Instruction/ maintenance manual of the product ATX612 Industrial Scientific
Go to page of 17
P/N 1706-3421 REV 03 Printed 0900-2000 Specifications Subject to Change 1001 Oakdale Road, Oakdale, P A 15071-1500 (412) 788-4353 • T oll Free 1-800-DETECTS F AX 412-788-8353 • Service Dept. 1-888-788-4353 Multi-Gas Monitor AT X 612 Instruction Manual GUARANTEED.
OUR MISSION Design - Manufactur e - Sell: Highest quality pr oducts for the pr eservation of life and pr operty . Pr ovide: Best customer service available.
Dear V alued Customer , Thank you for buying and using Industrial Scientific’ s Model A TX612 Multi-Gas Monitor . Y our A TX612 can be relied upon for dependable service, day after day . It has been designed, manufactured, tested and proven under the most scrutinizing conditions possible.
Multi-Gas Monitor AT X 612 1. W ARNINGS AND C AUTIONAR Y S T A TEMENTS 3 2. U NP ACKING T HE I NSTRUMENT 4 3. A TX612 F EA TURES 4 4. I NSTRUMENT O PERA TION 5 4 .1 Charging the Battery 5 4.1.1 Alkaline Battery Option 8 4 .2 Turning the A TX612 On and Off 8 4.
4 5 2. U NP ACKING THE I NSTRUMENT The shipping box should contain the following items. Account for each item before discarding the box. Q UANTITY P ART N UMBER D ESCRIPTION 1 1810-2707 A TX612 Multi-.
INDUSTRIAL SCIENTIFIC ATX612 OXYGEN LEL E BATT PPM % ON/OFF MODE 6 7 Hidden (+) Key Alarm Speaker On/Off/Mode Key Ultra-bright V isual Alarm Light Bar Battery Status Indicator Hidden (-) Key Display E.
8 9 CO H2S 02 LEL ATX612 VER 1.0 5 O 0 21.0 0 The remote sample tube should be connected to the barbed inlet fitting on the bottom of the A TX612 (See Figure 3). NOTE: When drawing a r emote sample, allow 2 seconds per foot (0.3 meter) of tubing length in addition to the normal sensor r esponse time befor e observing the instrument r eadings.
10 11 4.6.2 HIGH ALARM When a monitored gas concentration reaches the high alarm level setpoint, the instrument emits a continuous dual tone alarm. As with the low alarm condition, the red alarm light bar and backlight will flash simultaneously with the displayed gas value.
12 13 ZERO PRESS ZEROING 21.0 02 CAL 4.6.6 BA TTERY F AILURE When the battery has insufficient char ge to operate the instrument, “BA TTER Y F AIL” is displayed. The visual alarm will be activated and the instrument will emit a short beep once every second for approximately 30 seconds after which the instrument will turn itself off.
14 15 CODE 0 PRESS • T o continue full calibration, press (E) and the instrument will display the first sensor to be calibrated along with the calibration gas setting. The message APPL Y CAL GAS will scroll across the bottom of the display . NOTE: If the gas concentration does not match the setting, pr ess the MODE switch to abort calibration.
16 17 CODE PRESS CODE 123 PRESS • Press (+) to step to the INST ANT function and set instantaneous alarm values. The display will indicate INST ANT along with the scrolling prompt PRESS (+) FOR NEXT (E) TO SELECT . • Press (E) to enter the instantaneous alarms function.
18 19 The datalogging section of the instrument is always powered and a battery backup circuit protects it from loss of data for up to 40 minutes during battery changes. NOTE: The instrument must be stor ed on a battery char ger when not in use to pr event loss of data due to battery dischar ge.
20 21 • Remove the battery pack from the instrument as described in Section 8.2. • Remove the two screws which hold the battery module in place as shown in Figure 4. • Lift the Nicad battery module from the battery pack. • Place the new Nicad module in the battery pack.
22 23 I TEM P ART N UMBER D ESCRIPTION (Q TY ) 1 1705-8868 Main PC Board (1) 2 1705-6755 Display PC Board (1) 3 1705-9247 Interface PC Board (1) 4 1705-9031 Chassis (1) 5 1706-3447 Pump Assembly (1) 6.
24 25.
27 26 11 . D EF AUL T A LARM S ETTINGS GAS LOW ALARM HIGH ALARM O 2 19.5% 23.5% LEL 10% 20% CH 4 1.0% 1.5% CO 35 PPM 70 PPM H 2 S 10 PPM 20 PPM SO 2 2.0 PM 4.0 PPM NO 2 3.0 PPM 6.0 PPM Cl 2 0.5 PPM 1.0 PPM ClO 2 0.3 PPM 1.0 PPM NOTE: Factory alarm settings may not coincide with local r egulations.
28 29 Industrial Scientific portable gas monitoring instruments are warranted to be free from defects in material and workmanship for as long as the instrument is in service.
An important point after buying a device Industrial Scientific ATX612 (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Industrial Scientific ATX612 yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Industrial Scientific ATX612 - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Industrial Scientific ATX612 you will learn all the available features of the product, as well as information on its operation. The information that you get Industrial Scientific ATX612 will certainly help you make a decision on the purchase.
If you already are a holder of Industrial Scientific ATX612, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Industrial Scientific ATX612.
However, one of the most important roles played by the user manual is to help in solving problems with Industrial Scientific ATX612. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Industrial Scientific ATX612 along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center