Labconco XPert Filtered Balance System/Station Manual Warranty Labconco provides a warranty on all parts and factory workmanship. The warranty includes areas of defective material and workmanship, provided such defect results from normal and proper use of the equipment. The warranty for all Labconco products will expire one year from date of installation or two years from date of shipment from Labconco, whichever is sooner, except the following:
Part #3905501, Rev. B

TABLE OF CONTENTS CHAPTER 1: INTRODUCTION 1 About This Manual 3 Typographical Conventions 4 CHAPTER 2: PREREQUISITES 5 Support, Vibration & Movement Requirements 6 Temperature Variation Requirements 6 Humidity and Static Electricity Requirements 7 Background on Electrostatics or Static Electricity 7 Location and Air Current Requirements 8 Exhaust and Blower Requirements 8 Electrical Requirements 10 Space Requirements 10 CHAPTER 3: GETTING STARTED 11 Unpacking Your Enclosure 12 Installing the Filtered Enclosure on a Supporting Structure and Work Surface 12 Connecting to the Exhaust System (Optional) 15 Installation of HEPA Filters and Accessory Odor Control Carbon Filters 18 Connect the Electrical Supply Source to Filtered Enclosure 19 Set the Face Velocity with Speed Control Adjustment 21 Validating the Vented Enclosure 21 Sealing the Filtered Enclosure to the Work Surface 22 CHAPTER 4: HIGH PERFORMANCE FEATURES AND
SAFETY PRECAUTIONS 23 Components 25 Safety Precautions 31 CHAPTER 5: APPROPRIATE APPLICATIONS FOR YOUR
FILTERED ENCLOSURE 35 Routine Daily Work Procedures 35 Suitable Applications 37 HEPA Filter Applications, Suitability & Guidelines 37 Odor Control Carbon Filter Applications 38 Definition of Terms 39 Appropriate Chemicals for Odor Control Carbon Filters 40 Hazardous Misapplications for Odor Control Carbon

Filters with Volatile Chemicals 40 Chemical Carcinogen Use with Odor Control Carbon Filters 41 Prohibited Acid Use 41 CHAPTER 6 MAINTAINING YOUR FILTERED ENCLOSURE 43 Routine Maintenance Schedule 44 Decontamination 45 Determination of when to Replace HEPA Filters 45 How to Install a New HEPA Filter with Bag-In/Bag-Out Procedure for XPert Filtered Balance Systems and Stations 46 HEPA Filter Leak Test 47 Setting the Inflow Face Velocity with the Speed Control Adjustment 49 Calibrate and Operate the Airflow Monitors 50 Guardian Digital 1000 Airflow Monitor 52 Determination of When to Replace Odor Control Carbon Filters and How to Replace 56 Calculating Odor Control Carbon Filter Life 58 Initial Certification 59 Re-Certification 59 Fluorescent Light Replacement 60 Motorized Impeller Replacement 60 Speed Control Replacement 62 CHAPTER 7 ACCESSORIZING & MODIFYING YOUR
FILTERED ENCLOSURE 63 CHAPTER 8 TROUBLESHOOTING & SERVICER OPERATING LOG 69 APPENDIX A: FILTERED ENCLOSURE COMPONENTS AND REPLACEMENT PARTS 73 APPENDIX B: DIMENSIONS AND EXHAUST OPTIONS 79 APPENDIX C: FILTERED ENCLOSURE SPECIFICATIONS 83 APPENDIX D: QUICK CHART FOR THE FILTERED

CHAPTER 1 INTRODUCTION Congratulations on your purchase of a Labconco XPert™ Filtered Balance System or XPert™ Filtered Balance Station. Your enclosure provides personnel protection through superior containment while conserving energy at OSHA approved velocities as low as 60 feet per minute. It is the result of Labconco’s more than 50 years experience in manufacturing fume hoods and more than 30 years experience in manufacturing filtered enclosures. These enclosures will effectively contain toxic or noxious particulates when properly installed and operated. Each enclosure uses a single HEPA filter, which is rated at least 99.99% efficient for 0.3-micron particles. Additionally, an accessory Odor Control carbon filter may be added to adsorb nuisance odors from organic vapors, formaldehyde or ammonia and amines. The XPert filtered enclosures offer many unique features to enhance safety, performance, and energy savings. To take full advantage of them, please acquaint yourself with this manual and keep it handy for future reference. If you are unfamiliar with how high performance HEPA filtered enclosures operate, please review Chapter 4: High Performance Features and Safety Precautions before you begin working in the enclosure. Even if you are an experienced user, please review Chapter 5: Using Your HEPA Filtered Enclosure, which describes the XPert features so that you can use the filtered enclosure efficiently. For weighing hazardous and nuisance powders, Labconco recommends the purchase of the XPert Filtered Balance System or XPert Filtered Balance Station. Both XPert

filtered enclosures include a true bag-in/bag-out HEPA filter to properly protect personnel during filter changing operations. See chart below and contact Labconco for additional ordering information.
* It is recommended that a Labconco product specialist review the chemical application to determine if it is suitable. Consult Chapter 5 and Chapter 6.

This manual is designed to help you learn how to install, use, and maintain your filtered enclosure. Instructions for installing optional equipment or accessory carbon filters on your filtered enclosure are also included. Chapter 1: Introduction provides a brief overview of the filtered enclosure, explains the organization of the manual, and defines the typographical conventions used in the manual. Chapter 2: Prerequisites explains what you need to do to prepare your site before you install the filtered enclosure. Electrical and service requirements are discussed. Chapter 3: Getting Started contains the information you need to properly unpack, inspect, install, and certify the filtered enclosure. Chapter 4: High Performance Features and Safety Precautions explains how the XPert filtered enclosure operates and the appropriate precautions you should take when using it. Chapter 5: Using Your Filtered Enclosure discusses the basic operation of how to prepare, use and shut down your filtered enclosure. Chapter 6: Maintaining Your Filtered Enclosure explains how to perform routine maintenance on the filtered enclosure. Chapter 7: Accessorizing Your Filtered Enclosure explains acceptable modifications to the filtered enclosure or how to add accessories. Chapter 8: Troubleshooting contains a table of problems you may encounter while using the filtered enclosure including the probable causes of the problems and suggested corrective actions. Appendix A: Components and Replacement Parts contains labeled diagrams of all of the components of the filtered enclosures. Appendix B: Dimensions and Exhaust Options contains comprehensive diagrams showing all of the dimensions for the filtered enclosures.

Appendix C: Specifications contains the electrical requirements for filtered enclosures. Wiring diagrams are also included. Appendix D: Quick Chart provides an airflow table and test data for the filtered enclosures. Appendix E: References lists the various resources available that address laboratory ventilation and biosafety.
☞

CHAPTER 2 PREREQUISITES Before you install the filtered enclosure, you need to prepare your site for installation. You must be certain that the area is level and of solid construction. In addition, a dedicated source of electrical power should be located near the installation site to power the filtered enclosure, balance and other apparatus. Additionally, the enclosure should be strategically placed in the lab to provide efficient workflow. Carefully read this chapter to learn the requirements for your installation site:


Electrostatics can be troublesome in a balance enclosure. It is important to understand and, to the extent possible, control static charges. An electrostatic charged vessel, sample or enclosure can apply forces and lead to errors in weighing. The repulsion or attraction can be detected with micro, semi micro and analytical balances. Static charges can also lead to particulates being attracted to surfaces within the balance enclosure. Containment of harmful powders, prevention of cross-contamination and clean up are enhanced when static attraction of powders is minimized. The construction of the XPert Filtered Balance System and XPert Filtered Balance Station avoid the use of plastics, which are highly insulative. The advantages to the glass and epoxy-coated metal construction are twofold:
Electrostatic charges on a surface such as the wall of a balance enclosure are not created by moving air. Gases do not cause the charge. Impurities within the air impinging upon surfaces dictate the polarity and magnitude of the charge. The process, triboelectrification, occurs when the dust particles contact the surface, creating friction and electrons move across the interface.

The ability of a material to become polarized is a property known as permittivity. On highly insulative materials like acrylic, ions or charged molecules are strongly bound to the surface by polarization forces. The higher the force, the higher is the permittivity value of the material. It is suggested by balance manufacturers that the use of high permittivity materials, such as plastic be avoided. Since static electricity is a surface phenomenon, materials can also be classified by their surface resistivity measured in ohms per square. The table below lists the surface resistivity of various classes of material.
Surface Resistivity Table Material Surface Resistivity Example Conductive 0 → 105 Ω per square Skin, Metals Static dissipative 105 → 109 Ω per square Glass Antistatic 109 → 1012 Ω per square Polyethylene bag Insulative 1012 Ω per square→ Acrylic, Packing foam, Styrofoam Location and Air Current Requirements
The XPert Filtered Balance System uses an integral motorized impeller to draw room air past the operator and through the enclosure. This contaminated air is then pushed through the HEPA filter. The HEPA-filtered exhaust air is then forced out the top of the enclosure. An optional carbon filter may be installed on the downstream side of the HEPA filter to protect against nuisance odors.

The HEPA-filtered exhaust air can be recirculated into the laboratory or exhausted outside with the addition of the exhaust connection kit and remote blower listed in Chapter 7. The XPert Filtered Balance Station does not include the motorized impeller and must be exhausted to the outside by a remote blower. The XPert Filtered Balance Station remote blower may be switched from the blower switch on the enclosure. See electrical Requirements in Chapter 2, and Electrical Supply Connections in Chapter 3. Only one 6" exhaust connection is required for the 2', 3', or 4' XPert Filtered Balance Stations to exhaust to the outside. Data for the exhaust volume, noise pressure and enclosure static pressure loss are listed for each filtered enclosure model at face velocities of 60, 75, 80, 90, 100, and 105 fpm.
XPert Station Initial Static
Max. Equivalent
Pressure Loss with HEPA filter
Max. External
Resistance of 6" Duct
Static Pressure 2' 2' XPert Filtered Balance System 2' XPert Filtered Balance Station
(in w.g.)
.20" .19" .18" .16" .13" .12" 3' 3' XPert Filtered Balance System 3' XPert Filtered Balance Station
60 75 80 90 100 105
85 110 115 130 145 155
.35"- .41" .45" - .54" .48" - .57" .56" - .65" .62" - .72" .65" - .76"
48-53 49-55 50-56 51-57 53-58 54-59
400 220 200 140 80 70
.15" .12" .11" .09" .08" .07" 4' 4' XPert Filtered Balance System 4' XPert Filtered Balance Station
60 75 80 90 100 105
130 165 175 200 220 230
.37" - .43" .50" - .58" .53" - .62" .61" - .71" .68" - .79" .71" - .88"
48-53 52-56 53-57 55-58 58-61 60-64
130 65 50 35 25 20
.20" .19" .18" .16" .13" .12"
60 75 80 90 100 105
175 220 235 265 295 310
.40" - .46" .53" - .62" .57" - .66" .68" - .78" .76" - .87" .80" - .91"
49-54 54-58 58-61 60-63 62-66 64-67
100 60 50 35 25 20

Proper blower selection can be determined from these exhaust requirements and the total system static pressure loss. For outside exhaust, the enclosure must be connected to either a dedicated blower or a house exhaust system. Labconco offers accessory remote blowers listed in Chapter 7. Contact Labconco for blower sizing assistance. If the enclosure is connected directly to a house exhaust system, an adjustable damper (or valve) must be installed to control the airflow properly. This is equally important when a house exhaust system is controlling multiple filtered enclosures. See Chapter 7 for accessory adjustable damper ordering information. Electrical Requirements
Standard duplex electrical receptacles should be nearby for connecting the filtered enclosure, or other equipment, such as a balance for weighing operations. The enclosures include iris pass- throughs to allow electrical cords through the back of the enclosure without leaving a large hole for contaminants to escape. The remote blower for the XPert Filtered Balance Station may be connected to the blower switch on the enclosure. The blower switch is connected to a relay whose wires are terminated in the junction box behind the front panel. We recommend a maximum amperage of 6 amps for this circuit to the remote blower. Please refer to the wiring diagrams in Appendix C. Connect the blower wires inside the junction box to the remote blower per local electrical codes.
Space Requirements
The dimensions for the different models are shown in Appendix B: Dimensions.

CHAPTER 3 GETTING STARTED Now that the site for your filtered enclosure is properly prepared, you are ready to unpack, inspect, install, and validate your system. Read this chapter to learn how to:
Each enclosure model weighs between 125 to 195 lbs. each (55 to 85 kg). The shipping container allows for lifting with a mechanical lift truck or floor jack. If you must lift the enclosure manually, follow safe-lifting guidelines. Do not lift by the front air foil.

Carefully remove the shrink-wrap or carton on the enclosure and inspect it for damage that may have occurred in transit. If damaged, notify the delivery carrier immediately and retain the entire shipment intact for inspection by the carrier.
The United States Interstate Commerce Commission rules require that claims be filed with the delivery carrier within fifteen (15) days of delivery.
DO NOT RETURN GOODS WITHOUT THE PRIOR AUTHORIZATION OF LABCONCO. UNAUTHORIZED RETURNS WILL NOT BE ACCEPTED. IF ENCLOSURE WAS DAMAGED IN TRANSIT, YOU MUST FILE A CLAIM DIRECTLY WITH THE FREIGHT CARRIER. LABCONCO CORPORATION AND ITS DEALERS ARE NOT RESPONSIBLE FOR SHIPPING DAMAGES. Do not discard the packing material until you have checked all of the components and tested the enclosure. We recommend that you do not remove the enclosure from its shipping container until it is ready to be placed into its final location. Move the unit by placing a flat, low dolly under the shipping skid, or by using a floor jack.
☞
!
Do not move the enclosure by tilting it onto a hand truck. Installing the Filtered Enclosure on a Supporting Structure and Work Surface
Use caution when lifting or moving the enclosure. When installing the enclosure onto a chemical-resistant work surface or benchtop, ensure that the structure can safely support the combined weight of the enclosure and any related equipment. The work surface should be at least as wide as the enclosure to properly support it. The front of the enclosure should be aligned within

0.36" of the front of the work surface. Mounting holes are provided in the Labconco accessory work surfaces to secure the enclosure. Work Surface Specifications The work surface should be smooth, rigid, and durable, such as a chemical-resistant epoxy resin. The surface should be non-porous and resistant to the powders, solvents and chemicals used in conjunction with the XPert Filtered Enclosure. The work surface should also contain a dished recessed area for containing primary spills. Work Surface and HEPA Filtered Enclosure Installation
1. Level the base cabinets and the work surface. Work surface should be placed flush with the front of the base cabinet as shown in Figure 3-1. 2. Position the work surface in its intended location and with the front of the work surface towards you. (Rear mounting holes are located close to the rear edge.) 3. Secure the work surface to the base cabinet with a structural adhesive or silicone sealant. 4. Insert the supplied mounting screws in the four holes. Allow a minimum of 1/8" clearance under the head of the screw for positioning the enclosure. 5. Place the enclosure on the work surface and slide the rear flange and front air foil flanges under the mounting screw heads. 6. Tighten the four screws to complete the installation.


Figure 3-1 Filtered Enclosure Installation

!
The exhaust system should be installed by a qualified HVAC contractor.
The exhaust connection on the XPert Filtered Balance Station has been designed to accept 6" diameter ductwork as shown in Figure 3-2. The exhaust connection ships uninstalled on the top and needs to be fastened with the screws provided. See Chapter 7 for ordering exhaust transition kits for the XPert Filtered Balance System. Exhaust transition kits aid in the removal of chemicals or applications where a higher degree of powder and particulate removal is required. Review Chapter 2 for exhaust prerequisites and review Chapter 7 for ordering blower exhaust equipment. See Figure 3-2 for the XPert Station and Figure 3-3 for exhaust kit options for the XPert Filtered Balance System. Consult Labconco Customer Service should you require help sizing your blower for the exhaust volume and system static pressure loss. To ensure compatibility, the selected exhaust duct material should match the enclosure, procedures and chemical applications.

Figure 3-2 XPert Filtered Balance Station only Model Series 3940200, 3940300, and 3940400



Figure 3-3 Exhaust Duct Connection Kit shown installed on
XPert Filtered Balance System (Optional) Kit #3924400 (2'), 3924401 (3'), and 3924402 (4')

HEPA Filters HEPA filters are shipped installed with the gasket on the downstream side. The HEPA filters are leak checked at Labconco. A second leak check is recommended before using the enclosure and at least annually thereafter. Consult your Safety Officer and Chapter 6 for the HEPA Filter Leak Test and HEPA Filter Bag- In/Bag-Out Procedure. See Figure 4-2 for HEPA filter location, HEPA filter gasket, filter clamp bolts, and bag-in/bag-out flange. The XPert enclosures have a true bag-in/bag-out HEPA filter to safely dispose of the filter. All seams upstream of the HEPA filter are jacketed by negative pressure. This intrinsically safe design ensures containment. Carbon Filters (Accessory for Odor Control only) For carbon filter installation, first remove the upper perforated cover, secured by Phillips head screws, on the top of the enclosure. See Figures 4-1 and 4-2 for location of the upper perforated cover. See Chapter 7 for a list of available carbon filters. All carbon filters are simply gravity installed with the gasket side down and down stream on the exhaust side of the HEPA filter. Install the correct carbon filter for the application.
CARBON

Filter Type Appropriate Use HEPA 2' 3707900 3' 3707901 4' 3707902
HEPA filters are high-efficiency particulate air filters having a particulate removable efficiency of 99.99% for particles with a diameter of 0.3 micron. Organic Vapor Activated Carbon
Adsorbs organic compounds designated by NIOSH guidelines as acceptable for use with chemical cartridge-type respirators. Concentrations in the enclosure’s work area must not exceed the IDLH for the chemical and the exhaust from the enclosure must not exceed the TWA. Formaldehyde
2' 3937200 3' 3937300 4' 3937400
Formaldehyde requires the use of an impregnated carbon for the treatment of formaldehyde. Concentrations in the enclosure’s work area must not exceed the IDLH for the chemical and the exhaust from the enclosure must not exceed the TWA. Ammonia and
(Formalin) 2' 3937201 3' 3937301 4' 3937401
Treats ammonia, low molecular weight amines and other bases designated by NIOSH as acceptable for use with ammonia cartridge type respirators. Requires the use of an impregnated carbon. Concentrations in the enclosure’s work area must not exceed the IDLH for the chemical and the exhaust from the enclosure must not exceed the TWA. Important: Carbon filters do not provide any particulate protection, but provide odor control.
Amines 2' 3937202 3' 3937302 4' 3937402
Connecting the Electrical Supply Source to the Filtered Enclosure
115V Models Simply connect the 115V power cord supplied to the IEC electrical supply plug on the back of the enclosure. If using at 50 Hz operations, blower performance maximum airflow will be reduced by 17%. For the remote blower used for XPert Stations, follow the wiring diagram in Appendix C and wire the blower wires in the electrical junction box located behind the front panel. These wires are switched by the electrical relay via the blower switch.
XPert Station

The maximum circuit load for the electrical relay is 6 amps. Only the remote blower wiring is connected inside the electrical junction box. The lights and airflow monitor are powered from a standard duplex electrical receptacle located nearby for connecting the power cord. All wiring for the XPert Filtered Balance Station should be performed by a licensed electrician and conform to local codes. In most cases, the XPert Filtered Balance Station will require the use of shielded conduit to protect the wiring. See Figure 3-4. The grounding connection shall not be made to the terminal box cover. The building electrical supply system for the remote blower should include overload protection such as a switch or circuit breaker in close proximity and within easy reach of the operator. The switch or circuit breaker shall be marked as the disconnecting device. Consult NEC-2002 for proper installation.

Figure 3-4 XPert Filtered Balance Station Conduit Location for Wiring
External Remote Blower.

230V Models The same procedure applies for the 230V except it is shipped without a plug. Install the appropriate plug for your electrical specifications per local codes. Set the Face Velocity with the Speed Control Adjustment
For XPert Filtered Balance Systems, adjustment of the speed control gives the correct face velocity and is located behind the front panel. The face velocity should be adjusted from 60 to 100 fpm for weighing operations. (Consult your Safety Officer for airflow recommendations for your application). Containment is maximized at a setting within this range. Working at the lowest face velocity appropriate for the application will give the quietest operation. Face velocity measurements are made using an anemometer. An electric anemometer can be obtained from your laboratory supply dealer. Face velocity measurements should be taken in accordance with the Industrial Ventilation Manual. Using a small Phillips screwdriver, adjust the speed control to give the required face velocity. See Figures 6-1 and 6-2 in Chapter 6 to locate the speed control. The face velocity is increased by turning the speed control counter clockwise and clockwise to decrease face velocity.
Validating the Vented Enclosure
To determine the actual face velocity at the sash opening, airflow velocity readings are taken. This should be done across the sash opening of the enclosure in accordance with the Industrial Ventilation Manual. (See Appendix E) The “average face velocity” is achieved by taking readings in two rows across the enclosure with the readings 6" from the ends and evenly spaced every 12"; the first row is 3" down from the upper sash foil and the second row is 3" up from the work surface. A total of four readings will be taken for the 2' enclosure, six readings taken for the 3' enclosure and eight readings taken for the 4' enclosure and then averaged. Refer to Chapter 2 for proper airflow volumes for your particular model. The XPert enclosures have been tested at Labconco’s airflow test facility per ASHRAE 110-1995. All enclosures achieve an “as manufactured rating” of less than 0.05 part per million (ppm) at 4 liters per minute (lpm); AM

individual ratings). For “field use” ASHRAE testing contact Labconco for a certified on-site contractor. For particulate powder validation, Labconco had containment testing performed to validate the enclosures for sodium naproxen powders. The XPert HEPA filtered enclosures demonstrated excellent containment when used by an operator using excellent technique and good containment when used by an operator using marginal technique. While no enclosure can compensate for improper technique, these tests confirm that the XPert HEPA Filtered Enclosures provide a safe working environment. Labconco also performed extensive ASHRAE testing to validate the enclosures for gas containment. For copies of these validation reports, contact Labconco Customer Service. The XPert enclosures were also subjected to Biological Containment testing, using a modified Personnel Protection Test, as described in NSF/ANSI Standard Number 49. An aerosol challenge of approximately 5 x 108 spores of Bacillus subtilus var Niger were released in the enclosure and appropriate air sampling equipment established the number of spores that escaped through the sash opening. Each size enclosure was tested three times. All three sizes of the enclosure were tested at 75 FPM inflow and typically 1 or less than one spore escaped from the enclosure during each of the tests.
When the filtered enclosure has been set in place, ducted, (if necessary), and wired, it may be sealed at the work surface to prevent spilled materials from collecting under the walls. A bead of silicone sealant is recommended to seal the filtered enclosure to the work surface.

CHAPTER 4 HIGH PERFORMANCE FEATURES AND SAFETY PRECAUTIONS
High Performance Features: The patented* XPert Filtered Balance System and XPert Filtered Balance Station are designed to meet the needs of the laboratory scientist, and provide superior containment while conserving energy at OSHA approved “low flow” velocities as low as 60 feet per minute. The filtered enclosures have been tested to effectively contain toxic and noxious materials when properly installed and operated. What makes the filtered enclosures so unique is the revolutionary way they direct air into and through the contaminated air chamber. Labconco engineered the filtered enclosures to minimize the effects of turbulence. The containment-enhancing and aerodynamic designs of the upper sash foil, side air foils, lower air foil, upper dilution air supply, and rear perforated baffle all work in concert to produce horizontal airflow patterns that significantly reduce powder, chemical and particulate concentrations through the work area.
*(U.S. Patent No. 6,461,233)

These concentrations of materials are predominantly removed on the “first pass” of airflow through the chamber resulting in high performance containment.
The plenum and the HEPA filter are jacketed by negative pressure. Should a leak occur in the filter gasket or the plenum, the contaminated air is recaptured and refiltered. The XPert Filtered Balance System and XPert Filtered Balance Station include a true bag-in/bag-out HEPA filter disposal system to protect the worker from contact with hazardous powders and particulates. User’s of both products are encouraged to routinely check airflow with the use of the airflow monitor.

Not Shown 7, 9, 10, 11, 12, 13, 20, 21, 22, 23, 24, 25, 28 (See Figure 4-2)

Figure 4-1


Figure 4-2 HEPA Filtered Enclosure Airflow Diagram



Figure 4-3 · Figure 4-4


Monitor also displays a face velocity value, provides an RS232 output, a night setback mode and several auxiliary relay ports. Both airflow monitors are available options on all XPert models. See Figure 4-1 and 4-2.



16. Avoid the use of flammable gases or solvents in the enclosure if possible. Care must be taken to ensure against the concentration of flammable or explosive gases or vapors. Use of an open flame should be avoided in the enclosure. Open flames may disrupt the airflow patterns in the cabinet, burn the HEPA filter and damage the filter’s adhesive. Gases under high pressure should not be used in the enclosure as they may disrupt the airflow patterns of the cabinet.



CHAPTER 5 APPROPRIATE APPLICATIONS FOR YOUR FILTERED ENCLOSURE
Now that the installation of your filtered enclosure is completed, you are ready to use your filtered enclosure. Read this chapter to learn about:
Planning


NIOSH (National Institute for Occupational Safety and Health, USA) has established guidelines for chemical cartridge and HEPA filtered respirators. Suitable applications for the filtered enclosures are based on these guidelines. As with respirators, chemical contaminants are adsorbed or treated by carbon; particulates, powders and biohazardous materials are filtered by the HEPA filter before air is returned to the room. Biohazardous materials or other hazardous solids can be used with the HEPA filtered enclosure as long as appropriate decontamination occurs before filter changing operations. See Chapter 6 for decontamination procedures. Listed below are suitable applications for HEPA filters and accessory Odor Control Carbon Filters.
Shutdown

REFERENCES TO NIOSH OR OSHA GUIDELINES AND REGULATIONS APPLY TO ANY WORK PLACE UNDER THE JURISDICTION OF THE U.S. DEPARTMENT OF LABOR. OTHER COUNTRIES OUTSIDE THE U.S. HAVE ESTABLISHED STANDARDS, WHICH MAY DIFFER SLIGHTLY FROM THOSE USED AS GUIDELINES FOR THIS

PRODUCT. IT IS THE USER’S RESPONSIBILITY TO BECOME AWARE OF LOCAL REGULATIONS GOVERNING THE SAFE USE AND DISPOSAL OF CHEMICALS, CARBON AND HEPA FILTERS. KNOWLEDGE OF ESTABLISHED SAFE EXPOSURE LEVELS IS IMPERATIVE TO THE PROPER USE OF FILTERED ENCLOSURES.
NIOSH – National Institute for Occupational Safety and Health/Mine Safety and Health Administration. (U.S.A.) TWA – Recommended Exposure Limits expressed as a Time Weighted Average. The exposure limit for that chemical for up to a 10-hour workday, 40 hours a week. Expressed in units of parts per million or milligrams per cubic meter. Odor Threshold – The value in parts per million or milligrams per cubic meter for which one might expect to smell a chemical’s presence in the air. This value is very subjective and detection will vary with the sensitivity of one’s nose. The period of time until the odor threshold is reached in the exhaust stream can be estimated from Labconco exclusive computerized filter modeling program. Contact Labconco on carbon filter life for specific applications. See Chapter 6. Saturation Level or Time – There is a limit to the amount of chemical that can be adsorbed by activated carbon, or neutralized by chemically-treated carbon. Once the capacity of the carbon is reached, it is considered to be saturated and will adsorb (or neutralize) no further material; the outlet concentration of the chemical will equal the inlet concentration from that point until the filter is replaced. (Note that the capacity of activated carbon is not a constant, but varies with the inlet concentration). Labconco Technical specialists can determine with the computerized carbon- modeling program the estimated time saturation for a particular chemical. When using a HEPA filter alone in all filtered enclosures or in combination with an Odor Control carbon filter, the speed control will need to be increased to allow for HEPA filter loading. Measure filtered enclosure face velocity with an anemometer. IDLH (Immediately Dangerous to Life and Health). An atmosphere that poses an immediate hazard to life or produces

immediate irreversible health effects. IDLH concentrations should not be approached in the enclosure.
Below is a general set of rules to determine appropriateness of chemical usage.
Selected organic chemicals considered to be occupational carcinogens by NIOSH can be used in the filtered enclosure with carbon filters under rigid restrictions. See separate discussion on carcinogens for special instructions.
Organics must have time weighted exposure limits (TWA) of 1 PPM or greater.
Chemicals must have a detectable odor at concentrations below the TWA for the chemical.
Chemicals must be designated by NIOSH guidelines as acceptable for use with chemical cartridge-type respirators (the exception is formaldehyde and ammonia/amines, which used impregnated carbon). Chemicals not listed by NIOSH in the Pocket Guide must be approved by Labconco Product Specialist (or Engineering).
Inlet concentration must never exceed the IDLH (Immediately Dangerous to Life and Health) concentrations.
Chemicals having a recommendation by NIOSH of at least “Escape GMFOV” (Gas Mask Full-Face Respirator).
When evaporating a mixture of chemicals, the chemical having the lowest TWA will be used to determine if the mixture meets the guidelines.
Call a Labconco Product Specialist at 1-800-821-5525 for assistance in chemical appropriateness.
There is one scenario where the accessory carbon filter misapplication would be a part of a hazardous condition. If the

user continues to operate the enclosure with any of the following conditions present a potentially hazardous condition will exist:
Selected carcinogens may be used safely with Odor Control carbon filters under the following restrictions.
The use of a vented fume hood or ventilated enclosure with ducting to the outside is always the preferred method when working with carcinogens. The Odor Control carbon filters should only be used, as a last resort when venting to the outside is not an option. Another option is the use of the XPert Filtered Balance Station, which uses a remote blower. The potential carcinogens are listed in the NIOSH Pocket Guide to Chemical Hazards as “Ca.” Each potential carcinogen must have a TWA of 1 or greater; have minimum respirator recommendation of Escape GMFOV, and an odor threshold significantly lower than the TWA for the chemical.
The inlet concentration or the evaporation rate of the chemical must never exceed the TWA for the chemical. Consult a Labconco Technical Specialist for estimated saturation life. See Chapter 6 for an example of estimating saturation life. Another source is the Labconco chemical guide for carbon filtered enclosures.
The XPert Filtered Balance System motorized impeller cannot be exposed to acids. Where applications require the use of acids, a

separate ventilated enclosure or vented fume hood should be used with a remote blower ducted to the outside. No exceptions are permitted, as the impeller life span will be limited with acid use. NOTE: Low level acids can be used with the XPert Filtered
Balance Station, which uses a remote blower.

CHAPTER 6 MAINTAINING YOUR FILTERED ENCLOSURE Monitoring airflow and changing the filters is the primary maintenance required. Decontamination may be required and is reviewed in Chapter 6. Certification and recertification is reviewed in Chapter 6. Review this chapter on maintenance for the following: 1. Routine Maintenance. 2. Decontamination. 3. Determination of when to replace the HEPA filters. 4. How to install a new HEPA filter with the bag-in/bag-out procedure on XPert Filtered Balance Systems & Stations. 5. HEPA filter leak test. 6. Speed control adjustment and setting the inflow face velocity. 7. Operating and calibrating the airflow monitors. 8. Determination of when to replace Odor Control carbon filters and how to replace. 9. Calculating Odor Control carbon filter life. 10. Initial certification. 11. Re-certification. 12. Fluorescent light replacement. 13. Motorized impeller replacement. 14. Speed control replacement.

CARBON
CARBON
Weekly

The HEPA filters in the filtered enclosures gradually accumulate airborne particulate matter and powders from the enclosure and room. The rate of accumulation will depend upon the cleanliness of the room air, the amount of time the enclosure is operating, and the nature of work being done in the enclosure. In typical installations and usage, the HEPA filters will last two to five years before requiring replacement. Replace HEPA filters when face velocity drops below the recommended 60-80 fpm velocity, and the speed control is adjusted to full speed. Replace HEPA filters if it fails the HEPA Filter Leak Test in Chapter 6.
Annually

How to Install a New HEPA Filter with the Bag-In/Bag-Out Procedure for XPert Filtered Balance Systems and Stations
The XPert enclosures include a true bag-in/bag-out filter disposal system for safely removing the HEPA filter and protect the worker from toxic powders and particulates. See Figures 6-1 for the Bag- In/Bag-Out Procedure. For your safety, wear appropriate personal protective equipment during the change of HEPA filters. If working with biohazardous materials, use the decontamination procedure outlined in Chapter 6. (Consult your Safety Officer before performing filter changes).
1. With the enclosure OFF, remove the front panel secured by two screws on top of the enclosure. 2. Remove the bungee cords and retaining strap on the bag and extend it out. The retaining strap provides an extra level of safety even though the enclosure is under negative pressure when the blower is on. 3. Loosen the two filter clamp bolts with a 9/16" socket securing the HEPA filter. See Figure 4-1 and 4-2. 4. Place both arms in the rear gloves of the bag, and carefully draw the HEPA filter out of the enclosure. Rest the HEPA filter on the floor, table or cart. Remove arms from bag gloves. 5. Create two seals between the enclosure bag access port and the old HEPA filter. Seals can be heat sealed or sealed with tie wraps. Cut off the bag between the two seals and leave the bag stub on the access port to the enclosure. 6. Place the new HEPA filter in the new bag. Then place the new bag and new HEPA filter over the old bag stub and onto the enclosure access port. 7. Carefully remove the old bag stub using the single glove near the access port. Reattach the retaining strap on the perimeter of the new bag at the enclosure access port. Draw the old bag stub into the single glove and turn the glove inside out. Double seal this glove and cut off the glove with the contaminated bag stub inside. 8. Carefully install the new HEPA filter and uniformly tighten the two filter clamp bolts until the HEPA filter gasket is compressed 50%.


Figure 6-1 Bag-In-Bag-Out HEPA Filter Changing Procedure Diagram &
Filter Leak Test Diagram
Purpose After installing the new HEPA filter, the HEPA filter should be leak checked. This test is performed to determine the integrity of the HEPA filter, the filter housing, and the filter mounting frames. Leak testing is to be done by a qualified technician with calibrated equipment. Remove the top perforated exhaust cover by using a Phillips screwdriver to unfasten the (2) screws used to secure (2) clips. See Figure 6-1. The filter passes the leak test at .01% or better. Reference Leak Testing and Photometer scanning from the Institute of Environmental Services (IES-RP-CC001.3)

Apparatus 1. An aerosol photometer ATI model 2D, 2E, 2G or equivalent. Air Techniques Hamilton Associates inc. 11403 Cron Ridge Dr. Owings Mills, MD 21117 2. One aerosol generator of the Laskin nozzle(s) type. An aerosol of mineral oil or suitable liquid shall be created by flowing air through it. The compressed air supplied to the generator should be adjusted to a pressure of 10± 1 psig. during operation. Air Techniques Inc. Model TDA-4A or equal. One nozzle at 10 psig is (67.5 cfm x 100ug/l)/(Vol. of air), For the 2' at 90 fpm or 130 cfm, one nozzle @10 psig is 13,500/130 cfm = 52 ug/l. For the 3' at 90 fpm or 200 cfm, one nozzle @ 10 psig is 34 ug/l. For the 4' at 90 fpm or 265 cfm, one nozzle at 10 psig is 26 ug/l. 3. Mineral oil (Catalog #1491400). 4. Sampling Nozzle, Rectangular 1/2" x 3-1/4", Air Techniques, Inc. Procedure For the ATI 2G Photometer

11. Allow the generator to operate for a minimum of 15 seconds. For all integral motorized impeller models, scan the downstream exhaust side of the HEPA filter by passing the sampling nozzle of the gun in slightly overlapping strokes over the entire surface the filter, with the sampling port not more than 1 inch from the surface of the filter media. Scan the entire periphery of the filter and the gasket between the filter frame and the enclosure frame. Scanning shall be done at a traverse rate of not more than 2 inches per second.
sampling nozzle in the center of the remote blower exhaust. Acceptance Aerosol penetration shall not exceed 0.01 percent measured by the photometer Setting the Inflow Face Velocity with the Speed Control Adjustment
BALANCE
SYSTEM
1. Remove the front panel by loosening the (2) Phillips screws on top that secure the front panel. 2. The speed control is located on the electrical subassembly located behind the switched control panel and below the front panel. See Figure 6-1. 3. Adjust the speed control with a small Phillips screwdriver by turning the screw counterclockwise to increase blower speed or clockwise to decrease the blower speed. The speed control is very sensitive, so proceed with caution. 4. Measure the inflow velocity per the averaging technique outlined in Chapter 3 and adjust the speed control slowly for the desired speed. Allow the speed to stabilize and re- measure the inflow velocity to confirm. 5. Replace the front panel and tighten the screws.

Guardian Airflow Monitor (LED Monitor) Refer to Figure 6-2 for operation and calibration. Labconco Airflow Monitor / Airflow Switch Operation The Guardian Airflow Monitor (LED) consists of a circuit board and an airflow switch. This switch indicates airflow as safe or low. It does not provide an actual face velocity, but a small setscrew in the back of the sensor can adjust the airflow level that it classifies as “good/safe” or “low/alert.” The circuit board provides power to the sensor and also contains a “safe (green)” and “alert (red)” airflow LED indicators, as well as a “SILENCE ALARM” button to quiet the audio alarm. When first powered up, the PCB will light both red and green LED indicators and sound the alarm to indicate it is working. After 5 seconds, the air monitor will indicate either good or bad airflow based on what the connected airflow switch detects. For low airflow, the unit will wait for 10 seconds of bad indications before it sounds both the audio alarm and the red “alert” LED indicator. If the “SILENCE ALARM” button is pressed, the audio alarm will be silenced, but the red “alert” LED will remain on. The alarm is silenced indefinitely unless an airflow change is detected. If safe airflow is later detected for 10 seconds, the green “safe” LED will be lit and the “alert” (red) LED will be shut off. At any time the airflow is safe/good, one can press the SILENCE ALARM test button and the audio alarm and the red LED will turn on as long as this button is held down. The PCB has also a two-pin connector for use as an external output with isolated relay contacts that close when the red/alert LED is lit (low airflow). These relay contacts are not affected by the “SILENCE ALARM” button. The PCB is mounted behind the front panel using standoffs and an appropriate label is used to highlight the “SILENCE ALARM” button with clear areas for the red and green LED’s. No holes to allow sound to be broadcast louder are necessary. The PCB can be prepared as a factory special with an additional connector for the following external inputs, and having the following possible functions:

1. Ensure the flow switch and alarm circuit board are installed and operational. 2. Allow the enclosure to operate for at least two minutes. 3. If factory installed, the monitor will alarm at 60±10 fpm with the inflow velocity set at 90±10 fpm. 4. To change the factory setting, set the inflow velocity required by your Safety Officer to the desired alarm condition using the speed control adjustment procedure outlined in Chapter 6. 5. Once the alarm condition is set, use a small screwdriver to turn the adjustment screw on the airflow switch counterclockwise (facing the screw) until the “low” airflow red LED lights and the audible flow alarm sounds. 6. Adjust the inflow velocity to the nominal operating point required by your Safety Officer. 7. Over time the HEPA filter will load and eventually slow the inflow velocity. Once the alarm condition is met, simply increase the speed control outlined in Chapter 6 or replace the HEPA filter if the speed control is maximized. 8. The table below lists typical alarm conditions based on normal operating conditions. Typical alarm conditions are set at face velocities of 10 to 30 feet per minute below the normal operating conditions due to supply air and exhaust air fluctuations, as well as room air cross drafts. Consult your Safety Officer for proper operating speeds.
Alarm Condition Set Point
Speed 100 ± 10 fpm 70-90 fpm 90 ± 10 fpm 60-80 fpm 80 ± 10 fpm 50-70 fpm 70 ± 10 fpm 50-60 fpm 60 ± 10 fpm 50 fpm
Enclosure Operating In-Flow · Speed


Figure 6-2 Guardian Airflow Monitor (LED) with Airflow Switch
Refer to Figure 6-3 for operation and calibration. Refer to the separate Labconco 1000 Alarm Manual for detailed information.

Operation The Guardian Digital Airflow Monitor consists of the airflow sensor, the Alarm Unit, and the 15 VDC power supply. For 115V operation the alarm unit is powered by plugging the power supply into the factory-prepared digital airflow monitor socket. For 230V operation, the Alarm Unit is powered by plugging the power supply into a building outlet. The alarm has “Enter,” “+,” and “-” buttons to program the monitor. There is also a green LED “SAFE,” yellow LED “CAUTION,” and red LED “LOW” with audible alarm for airflow conditions. The audible alarm can be permanently muted if desired. The Guardian Digital 1000 Airflow Monitor displays a face velocity value, provides an RS232 communications port to a PC or building computer system, can be configured for external input connections such as night setback or external alarm and provides up to three output relays that can be configured. For complete detailed information, please refer to the separate Labconco 1000 Alarm User’s Manual provided with the enclosure. Digital 1000 Calibration The filtered enclosure comes factory calibrated with the monitor alarm at 60 ± 10 fpm with the inflow velocity set at 90± 10 fpm. To change the calibration, review the Labconco 1000 Alarm User’s Manual that comes with the enclosure. To successfully calibrate, it will be necessary to change the face velocity by adjusting the airflow exhaust volume with the speed control. The exhaust volume can be adjusted with the speed control or by using an adjustable damper for outside exhaust. NOTE: For accurate calibration, the front panel needs to be in place for each low calibration and high calibration set points to simulate real usage. Typical calibration conditions are set at face velocity air sample differences of a least 50 feet per minute. However, the airflow monitor is factory set and calibrated with a difference of at least 20 fpm and can be changed in the CAL CONFIG MENU by changing the “lower/higher air sample difference.” If changed, then also change the “sensor difference” in CAL CONFIG MENU from 10% to 3%. To change the red “Low” airflow alarm and yellow “CAUTION” adjust the values in CAL CONFIG MENU. Set “WARNING or CAUTION” air reset to 3 fpm. To complete the CAL CONFIGURATION, be sure to enter “Done.”

To start the calibration, use the Labconco 1000 Manual and enter the “CALIBRATIION” mode on the display. Follow the instructions on the display and alter the low and high exhaust volumes with the speed control. The following suggested inflow face velocity speeds are recommended to successfully calibrate. Typical low air alarms are set 10-30 fpm below operational speeds. Typical yellow cautions are set 3-10 fpm above the red low air alarm. Be sure to enter “Done” after completing the calibration in “SETUP.” Once completed, go to “RUN” and hit “ENTER.”
High Calibration Set Point 50 fpm 60 fpm 50-60 fpm 100-120 fpm 60-70 fpm 80 fpm 50-70 fpm 100-140 fpm 70-90 fpm 100 fpm 60-80 fpm 110-140 fpm Factory Settings 60 ± 10 fpm 90 ± 10 fpm 60 ± 10 fpm 90 ± 10 fpm
Nominal Enclosure Operating In Flow Speed
Low Calibration Set Point
Low Air Alarm Set Point


Figure 6-3 Guardian Digital 1000 Airflow Monitor

Determination of When to Replace Odor Control Carbon Filters and How to Replace
The carbon filters MUST be replaced when any one of the following two conditions are met:
There are four means of determining when its time to change the carbon filters (not shown in the order of preference).
Odor - A person’s sensitivity to odor, tolerance of odor and their comfort level under odoriferous conditions vary with the individual. While odor is an indicator that chemicals are passing through the carbon filter, several points need to be understood:
Detection Tubes - Color change indicators can be used to measure the concentration of the chemical at the exit side of the carbon filter or in the outlet exhaust. A kit including syringe pump and flexible tubing can be purchased as an accessory from Labconco (Catalog # 6924900). Labconco Customer Service Representatives are supplied with detector tube catalog numbers, as well as telephone numbers to direct you to where to purchase these items.
For Organic, Formaldehyde and Ammonia, chemical specific detector tubes should be purchased when installing fresh filters. Each kit contains instructions on how many strokes of the syringe are required to obtain the stated sensitivity. The sampling syringe is connected to the filtered enclosure exhaust. Connect the syringe to the detector tube while the system is running and pull the air through the tube with the syringe. Each stroke of the

syringe represents a 100-ml sample and corresponds to the number of strokes necessary to give the indicated color changes. Due to the wide variety of organics and varying TWA’s, it is recommended that specific detector tubes be purchased directly from Sensidyne, Draeger or your laboratory supply dealer. Alternate detector pumps can also be purchased from your laboratory supply dealers. The vast majority of detector tubes available start measuring at the TWA. When a user observes a color change in the tube, they should replace the filter immediately. If no detector tube for your specific chemical is available, other means of detection must be used.
Time - For applications that have very consistent inlet concentrations and operating time, “Time” can be used to anticipate saturation or TWA levels based on prior experience. However, this does not replace the need for sampling. Consult Labconco technical specialist for an estimate of carbon filter life based on chemical usage. Detector tubes, or analytical instrumentation should always be used to determine concentrations in the carbon filter. It is recommended that the carbon filters be checked with detector tubes or other means at intervals of 20% of the total estimated filter life. The exception to the 20% recommendation is formaldehyde and any carcinogen or suspected carcinogen. These more hazardous chemicals must be checked at least every 10% of the total estimated time.
Analytical Instrumentation - This is the most accurate means of measuring concentrations of any chemical. It is the method of choice when no detector tubes are available or the tubes are not sensitive enough to measure at the TWA concentration for the chemical. This method is also to be used to determine saturation when the chemical concentration is below the measurement range of detector tubes.
Odor Control Carbon Filter Replacement Procedure – See Figures 4-1 and 4-2 for Carbon Filter Replacement.

Labconco developed a modeling program to estimate the filter life for typical carbon filters. Since filter life is dependent on the chemical used, the airflow, filter size, and the dwell time, refer to the Chemical Guide for the Paramount® Filtered Enclosure. The estimated life for Odor Control carbon filters for the filtered enclosures is conservatively calculated at 50% or half of the published values for the Paramount in the Chemical Guide. For example, if you use isopropyl alcohol to disinfect and use approximately 100 ml per week during 2 hours of use per day then follow these steps to calculate the concentration in parts per million (ppm).
Steps for Calculating PPM and Filter Life 1. Determine the amount of the proposed chemical lost to evaporation over a given amount of time. For example, if you use isopropyl alcohol and lose approximately 100 ml per week during 2 hours of use per day. 2. Convert the amount lost into ml/min. For this example: 100 ml X 1 week X 10 hours = 100 ml lost week 10 hours use 600 minutes 600 minutes 3. Convert ml/min to PPM by multiplying ml/min by the conversion factor found in the second to the last column on the right. For isopropyl alcohol .17 x 41 = 7.0 PPM. 4. Find the PPM value on the chart that comes closest to the value you just calculated in step #3. In this example, round up to 10 PPM, which is close to the calculated 7.0. We may approximate the filter life to be around 155 hours of actual use, but use 50% of this for the Odor Control filters or 78 hours. 5. Insert the estimated filter life into the estimated usage to determine how long filters will last.
=.17 ml/min
78 hours filter life 10 hours per week use =7.8 weeks before filter saturation

Formaldehyde only For formaldehyde, use 10% of the impregnated carbon weight. Formalin is 37% formaldehyde by weight. The density of formalin is 1.08 g/ml.
Formalin 2' 18 x 18 x 1 7.0 0.7 lbs./318g 796 ml 3' 30 x 18 x 1 12.0 1.2 lbs./545g 1364 ml 4' 42 x 18 x 1 16.8 1.68 lbs./763g 1910 ml Ammonia only For ammonia, use 10% of the impregnated carbon weight. Assume use of a 50% solution of ammonia for these calculations.
Adsorbed Volume of 50%
II Carbon Adsorbed weight of
Solution 2' 18 x 18 x 1 9.2 0.92 lbs./418g 836 ml 3' 30 x 18 x 1 13.5 1.35 lbs./613g 1226 ml 4' 42 x 18 x 1 18.9 1.89 lbs./859g 1718 ml
The filtered enclosure has been certified at the factory for an inflow velocity of 90±10 fpm along with the HEPA Filter Leak Test. The filtered enclosure should be certified for the proper inflow velocity required by your Safety Officer. It is also a conservative recommendation to perform the HEPA Filter Leak Test again should there be any damage caused during transport.
In addition, the following tests should also be performed at the user’s discretion:

The motorized impeller must be replaced as a complete unit. When the motorized impeller is replaced, the capacitor may also be replaced. See Appendix A for Replacement Parts Diagram. See Figure 6-5 for an isometric view of the motorized impeller plenum assembly. The HEPA filter rests on top of the motorized impeller assembly.


Figure 6-4 Motorized Impeller Replacement


CHAPTER 7 ACCESSORIZING AND MODIFYING YOUR FILTERED ENCLOSURE
There are several ways to accessorize and modify the filtered enclosure for your individual requirements. These include the addition of accessory work surfaces, airflow monitors; exhaust transition kits, remote blowers, exhaust dampers, filters, storage cabinets and utility shelf kit.
An optional dished work surface is available to attach to the filtered enclosure. Dished work surfaces are contoured to fit the dimensions of the filtered enclosures to contain spills. Epoxy is chemical resistant. See Figure 3-1 for installation. Catalog # Description Dimensions (W x D x H) 3909900 Black, 2-foot wide 24" x 29" x 1" 3909901 Black, 3-foot wide 36" x 29" x 1" 3909902 Black, 4-foot wide 48" x 29" x 1" 3909903 Gray, 2-foot wide 24" x 29" x 1" 3909904 Gray, 3-foot wide 36" x 29" x 1" 3909905 Gray, 4-foot wide 48" x 29" x 1"

The Guardian Digital 1000 Airflow Monitor or Guardian LED Airflow Monitor allows you to continuously monitor airflow through the enclosure. Either monitor is usually provided with the XPert enclosures. Description Order Part #’s Individual Parts Required Guardian Airflow Monitor 115V or 230V
1 ea. 3811500 Airflow Monitor Printed Circuit Board 1 ea. 3910700 Airflow Sensor w/ Connector 2 ea. 1885405 #6-32 x .31" Phillips machine screw to mount Airflow Monitor Circuit Board 2 ea. 1891808 #6-32 x .50" Phillips thread cutting screw to mount Airflow Sensor Guardian Digital 1000 Airflow Monitor
115V 1 ea. 3908500 115V Guardian Digital Monitor, Airflow Sensor, Sensor Cable, and 15 VDC power supply 230V 1 ea. 3908501 230V Guardian Digital Monitor, Airflow Sensor, Sensor Cable, and 15 VDC power supply Add appropriate Front Panel with cutout to mount Digital 1000 Airflow Monitor 1 ea. 3904803 2' Front Panel w/ cutout or 1 ea. 3904804 3' Front Panel w/ cutout or 1 ea. 3904805 4' Front Panel w/ cutout NOTE: The cutout can also be made with a reciprocating saw and
a 1/8" diameter bit. The centerline of the cutout is located 3.3" from the left side and 4.2" from the bottom. The 1/8" mounting holes are spaced 4.38" apart top to bottom and centered over the cutout. The cutout dimensions are 2.43" wide by 4.13" tall.
Exhaust transition connects to the XPert Filtered Balance System with integral motorized impellers from the top so the duct can be routed up and outside the building. The transition is available for 6" PVC duct. The transition should be sealed to the top of the enclosure with silicone sealant to prevent leaks. See Figure 3-3.

Order the following: Part Description 3924400 2' Exhaust Transition Kit 3924401 3' Exhaust Transition Kit 3924402 4' Exhaust Transition Kit 4. Remote Blowers Has a 1/4 hp direct drive motor and corrosion-resistant epoxy- coated steel housing and wheel with blower inlet of 6.00" ID. Outlet dimensions are 4.25" x 7.38" OD.
CFM @ Static Pressure-Inches of H2O S.P. 0.0" 0.125" 0.25" 0.50" 0.75" 0.87" CFM 595 560 515 420 300 167 Catalog # Description Shipping Wt.
(lbs./kg.) 4863500 Remote Blower, 115 V, 60 Hz. 4.4 amps 35/16 4863501 Remote Blower, 115/230 V, 50 Hz, 5.6/2.8 amps 35/16
7053501 Explosion-Proof Remote Blower, 115 V, 60 Hz, 4.4 amps 40/18

Figure 7-1

Exhaust dampers allow adjustments required to maintain proper airflow for roof-mounted blowers or house exhaust systems. This would primarily be useful only for the XPert Filtered Balance Station, which does not have an integral blower, but does have a 6" nominal exhaust outlet. Catalog # Description 3924000 6" Epoxy Coated Steel In-Line adjustable damper 4724200 6" PVC In-line adjustable damper 6. Accessory Filters HEPA Filter HEPA Filter is 99.99% efficient on particles 0.3 micron.
Width Catalog # Size 2' 3707900 18" x 18" x 3.31" 3' 3707901 30" x 18" x 3.31" 4' 3707902 42" x 18" x 3.31" HEPA Filter Bag-In/Bag-Out Bag The bag helps contain hazardous particulate matter during filter changing operations.
Width Catalog # Size 2' 3776003 48" Circum. x 54" lg. 3' 3776004 73" Circum. x 96" lg. 4' 3776005 96" Circum. X 96" lg. Odor Control Carbon Filter Provides granular activated carbon or impregnated carbon used for Odor Control applications on integral motorized impeller models exhausted back into the room. Filter Classification and Size Catalog # X Pounds 2' Organic 18 x 18 x 1 3937200 5.5 lbs. activated 3' Organic 30 x 18 x 1 3937300 9.2 lbs. activated 4' Organic 42 x 18 x 1 3937400 12.9 lbs. activated 2' Formaldehyde 18 x 18 x 1 3937201 7.0 lbs impregnated 3' Formaldehyde 30 x 18 x 1 3937301 12.0 lbs. impregnated 4' Formaldehyde 42 x 18 x 1 3937401 16.8 lbs. impregnated 2' Ammonia & Amines 18 x 18 x 1 3937202 8.0 lbs. impregnated 3' Ammonia & Amines 30 x 18 x 1 3937302 13.5 lbs. impregnated 4' Ammonia & Amines 42 x 18 x 1 3937402 18.9 lbs. impregnated

SOLVENT ACID Size/Description Dual Doors Right Hinge Left Hinge Dual Doors Right Hinge Left Hinge 48" 9902000 - - 9901000 - - 36" 9902100 - - 9901100 - - 30" 9902200 - - 9901200 - - 24" - 9902300 9902400 - 9901300 9901500 18" - 9901400 9901600 12" - 48" w/Self-Closing Doors 9903000 - 36" w/Self-Closing Doors 9903100 - 30" w/Self-Closing Doors 9903200 - 24" w/Self-Closing Doors - 9903300 9903400 - - - 24" ADA - 9906000 9906100 - 9905000 9905200 24" ADA w/Self-Closing Doors
- 9906200 9906300 - - -
18" ADA - 9905100 9905300 12" ADA -
STANDARD BASE VACUUM PUMP Size/Description Dual Doors Right Hinge Left Hinge Dual Doors Right Hinge Left Hinge 48" 9900000 - 36" 9900100 - 30" 9900200 - 24" - 9900300 9900600 - - - 18" - 9900400 9900700 - 9907000 9907100 12" - 9900500 9900800 - - - 48" w/Self-Closing Doors - 36" w/Self-Closing Doors - 30" w/Self-Closing Doors - 24" w/Self-Closing Doors - 24" ADA - 9904000 9904300 - - - 24" ADA w/Self-Closing Doors
-
18" ADA - 9904100 9904400 - - - 12" ADA - 9904200 9904500 - - -



Refer to the following table if your filtered enclosure fails to operate properly. If the suggested corrective actions do not solve your problem, contact Labconco for additional assistance. PROBLEM CAUSE CORRECTIVE ACTION Contaminants outside of enclosure.
See “Certifying the Enclosure” Chapter 3 and “Safety Precautions” Chapter 4 sections in the manual. (Ref. Appendix D). Restriction of the baffle air slots or blockage of the exhaust outlet.
Improper user techniques for the enclosure.
Remove obstruction to ensure that all air slots and the exhaust outlet are unobstructed.
External factors are disrupting the enclosure airflow patterns or acting as a source of contamination.
See “Location Requirements” Chapter 2, “Certifying the Enclosure” Chapter 3, and “Safety Precautions” Chapter 4 sections of this manual. (Ref. Appendix D).
Enclosure has improper face velocity.
Have enclosure certified and check exhaust system. Check HEPA filters for loading. Adjust enclosure speed control. Enclosure should have an average face velocity of 60-100 fpm for weighing procedures and 75-105 fpm for biohazardous procedures.


Plug the enclosure into appropriate electrical service. Circuit breaker(s) or Ground Fault Interrupter.
Plug enclosure into appropriate electrical service. Circuit breaker(s) tripped.
Power cable to circuit board is disconnected and needs to be connected. Sensor cable needs to be connected. Power supply is not plugged into proper voltage; plug in power supply. Verify that all airflow monitor interface cables are connected. Check fuses on enclosure. No audible alarm. Alarm has been temporarily silenced using “SILENCE ALARM” or “enter” buttons.

PROBLEM CAUSE CORRECTIVE ACTION Airflow Monitor Malfunction (Continued)
Airflow monitor was not properly adjusted. Repeat calibration steps outlined in this manual in Chapter 6 or in the supplied Digital 1000 Monitor Manual. Constant audible alarm.
Wrong alarm set point.
Check airflow and calibration of airflow monitor. See Chapter 6. Continuous alarm. Check the face velocity of the enclosure as the airflow of the system may have changed. If incorrect, adjust the speed control to increase face velocity. The HEPA filter may have become loaded. If face velocity is correct, calibrate the airflow monitor outlined in this manual in Chapter 6. Monitor alarms; air way to airflow monitor sensor is blocked by insects, dust or debris.
Lightly clean the airway with clean air. Be careful not to touch sensitive electrical components.
Audible disable will not stay operational.
An alarm condition must be continuously present before the audible alarm can be silenced. If flow conditions fluctuate near the alarm set point, the airflow monitor will automatically reset it. Action should be taken to bring the enclosure airflow into proper operating parameters or adjust the alarm set point lower.

Customer Name Model Number
Date Installed Serial Number
Unit Location · Date Comments

APPENDIX A FILTERED ENCLOSURE COMPONENTS AND REPLACEMENT PARTS The components that are available for your filtered enclosure are listed. The parts shown are the most commonly requested. If other parts are required, please contact Product Service.

Item Qty. Part Number Description 1 1 3911000 Glass, Side 2A 1 3911900 Glass, Sash 2' 2B 1 3911901 Glass, Sash 3' 2C 1 3911902 Glass, Sash 4' 3A 1 3911800 Sash Assembly, 2' 3B 1 3911801 Sash Assembly, 3' 3C 1 3911802 Sash Assembly, 4' 4A 1 3905600 Lower Air Foil, 2' 4B 1 3905601 Lower Air Foil, 3' 4C 1 3905602 Lower Air Foil, 4' 5A 1 3907400 Bracket, Air Foil R.H. 5B 1 3907401 Bracket, Air Foil L.H. 6A 1 3907500 Stop, Sash R.H. 6B 1 3907501 Stop, Sash L.H. 7A 1 6916500 Latch, Sash 7B 1 3906700 Bracket, Latch 7C 1 1927403 Spring, Compression Latch 7D 2 1893206 Screw, 8-32 x .38" Lg. Phillips Pan S.S. Type F 8A 1 3776003 Bag, 2' – for Bag-In/Bag-Out on XPert 8B 1 3776004 Bag, 3' – for Bag-In/Bag-Out on XPert 8C 1 3776005 Bag, 4' – for Bag-In/Bag-Out on XPert 9A 2 1889316 Screw, 10-24 x 1.00 PH Phil S.S. 9B 2 1912108 Washer, Plastic .194 ID. x .380 OD 9C 2 7868402 Bushing, Spacer .31 x .63 10 1.9 Ft 6913700 Wiper, Sash Seal 11A 2 1936800 Bushing, Heyco closed 1.50 Dia. 11B 2 1934601 Bushing, Heyco w/ Flex Shutter 1.50 Dia. 12A 1 3915401 Side Air Foil 12B 2 1889912 Screw, #6-32 x .75" Oval Hd. Type F (S.S.) 13A 1 3907700 Reflector, Light 2' 13B 1 3907701 Reflector, Light 3' 13C 1 3907702 Reflector, Light 4' 13D 2 1889308 Screw, Mach. 10-24 x .50 PH Phillips 14A 1 3708300 Lamp, Fluorescent 2' Enclosure (F15T8-SP35) 14B 1 9721902 Lamp, Fluorescent 3' Enclosure (F17T8) 14C 1 9721901 Lamp, Fluorescent 4' Enclosure (F25T8) 16A 1 3911700 Bracket, Right Reflector Support 16B 1 3911701 Bracket, Left Reflector Support 16C 4 1893108 Screw, #8 x 1/2 AB Hex Washer Head

Item Qty. Part Number Description 17A 1 3918500 Wiring Harness Main, 115V 17B 1 3918600 Wiring Harness Main, 230V 18A 1 3811500 Guardian Airflow Monitor Printed Circuit Board 18B 2 1885405 Screw Mach #6-32 x .31 PH Phil. S.S. 19 1 3926800 Label, Front XPert 20A 1 3910700 Airflow Switch 20B 2 1891808 Screw, Thread Cutting #6-32 x .50 PH 21 Multiple 1905617 Nut, KEPS 10-24 22 2 1307000 Switch, Rocker (2 position) 23A 1 or 2 3904800 Front Panel, 2' 23B 1 3904801 Front Panel, 3' 23C 1 3904802 Front Panel, 4' 23D 1 3904803 Front Panel 2' with Digital Monitor Cutout 23E 1 3904804 Front Panel, 3' with Digital Monitor Cutout 23F 1 3904805 Front Panel, 4' with Digital Monitor Cutout 23G 2 4472400 Screw, Thread cut 10-24 x .50 Type F PH Phillips 24A 1 3916500 Impeller, Motorized 2' or 3' Enclosure, 115V 24B 1 3916501 Impeller, Motorized 2' or 3' Enclosure, 230V 24C 1 3903300 Impeller, Motorized 4', 115V 24D 1 3903301 Impeller, Motorized 4', 230V 24E 4 1601800 Vibration Isolator Mount 25A 1 1306900 Capacitor, 2', 115V 10 MFD 25B 1 1306900 Capacitor, 3', 115V 10 MFD 25C 1 1306600 Capacitor, 4', 115V 15 MFD 25D 2 1306800 Capacitor, 2', 230V 4 MFD (effective 2MFD) 25E 2 1306800 Capacitor, 3', 230V 4 MFD (effective 2MFD) 25F 1 1306800 Capacitor, 4', 230V 4 MFD 26A 1 1235400 Ballast, Fluorescent 2', 115V 26B 1 1235500 Ballast, Fluorescent 2', 230V 26C 1 1294000 Ballast, Fluorescent 3' or 4', 115V or 230V 28A 1 1325500 Speed Control, 115V 28B 1 1325501 Speed Control, 230V 29A 1 3910800 Upper Diffuser Screen 2' 29B 1 3910801 Upper Diffuser Screen 3' 29C 1 3910802 Upper Diffuser Screen 4' 30A 2 3916700 Clip, Upper Diffuser

Item Qty. Part Number Description 30B 2 1893708 Screw, #10 x .50 PH Phil. Thd. Forming 31A 2 1881196 Capscrew, 3/8 –16 x 6.0 Lg. Hx. Hd. 31B 2 1924403 Sealing Washer 3/8 32A 1 3707900 HEPA Filter (2') 18 x 18 x 3.32 w/ Gasket 32B 1 3707901 HEPA Filter (3') 30 x 18 x 3.32 w/ Gasket 32C 1 3707902 HEPA Filter (4') 42 x 18 x 3.32 w/ Gasket 33A 1 3937200 2' Odor Control Organic Vapor Carbon Filter 33B 1 3937201 2' Odor Control Formaldehyde Carbon Filter 33C 1 3937202 2' Odor Control Ammonia Carbon Filter 33D 1 3937300 3' Odor Control Organic Vapor Carbon Filter 33E 1 3937301 3' Odor Control Formaldehyde Carbon Filter 33F 1 3937302 3' Odor Control Ammonia Carbon Filter 33G 1 3937400 4' Odor Control Organic Vapor Carbon Filter 33H 1 3937401 4' Odor Control Formaldehyde Carbon Filter 33I 1 3937402 4' Odor Control Ammonia Carbon Filter 34 1 7907802 Digital Monitor Outlet, 115V only, 230V connected at laboratory outlet location. 35A 1 1333800 IEC Power Inlet 35B 1 1305800 Power Cord, 115V, Right Angle 35C 1 1334200 Power Cord, 230V, Right Angle 36A 1 - (115V) 1327200 Circuit Breaker, 3 AMP 36B 2 – 230V 1327205 Circuit Breaker, 2 AMP 37A 1 (115V) 3908500 Digital Airflow Monitor, 115V – 15VDC Adapter, Airflow Sensor 37B 1 (230V) 3908501 Digital Airflow Monitor, 230V – 15 VDC Adapter, Airflow Sensor 37C 1 1595621 Hole Plug, 1.187 Dia. 38A 1 3919800 Exhaust Duct, 6" (XPert Station only) 38B 4 1893708 Screw, #10 x .50 PH Phil. Type AB 39A 1 3919700 XPert Station Top Panel 2' 39B 1 3919701 XPert Station Top Panel 3' 39C 1 3919702 XPert Station Top Panel 4' 40A 1 1952500 Gauge, Pressure Mini Helic II 40B 1 3667400 Bracket, Gauge 40C 2 1905617 Nut, #10-24 KEPS 40D 2 1551700 Tubing, 3/16" ID x 5/16" OD, PBC 7.00" Lg. 40E 2, 1" Long 3788200 Snubber, Filter Pressure Gauge 41A 1 1289101 115V Relay (for Remote Blower on XPert Filtered Station) 41B 1 1289102 230V Relay (for Remote Blower on XPert Filtered Station) 42 1 3925600 Harness, Relay (for Remote Blower on XPert Filtered Station)

Use for XPert System and XPert Station

33 Placed on ledge
above HEPA · filter

Use for XPert System, XPert Station, Digital Monitor and Bag

42 Location only Not shown
39 Needs to be sealed on XPert Station
41, Location only Not shown
Ships uninstalled · as shown

See the following dimensions and exhaust options for all the XPert Filtered Balance Systems and XPert Filtered Balance Stations.


Figure B-1 XPert Filtered Balance System


Figure B-2 XPert Filtered Balance Station


Figure B-3 Exhaust Duct connection Kit (Optional Installation on Integral Blower Models
such as the XPert Filtered Balance System)











APPENDIX D QUICK CHART FOR THE FILTERED ENCLOSURES *excludes XPert Filtered Balance Station, since it has no integral blower. To determine the actual face velocity at the sash opening, airflow velocity readings will need to be taken. This should be done across the sash opening of the enclosure in accordance with the Industrial Ventilation Manual. (See Appendix E) The “average face velocity” is achieved by taking readings in two rows across the enclosure with the readings 6" from the ends and evenly spaced every 12"; the first row is 3" down from the upper sash foil and the second row is 3" up from the work surface. A total of four readings will be taken for the 2' enclosure, six readings taken for the 3' enclosure and eight readings taken for the 4' enclosure and then averaged.
Model Size 2' 3' 4' Sash Height from Work Surface (inches) 9.44 9.44 9.44 Total Open Area with Bypass (sq. ft.) 1.45 2.20 2.95 Exhaust Airflow Volume at 90fpm (CFM) 130 200 265 Alarm Airflow Volume at 60 fpm (CFM) 85 130 175 Initial Mag. Gauge Pressure Reading (inches H2O) 0.20 to .045 0.15 to 0.40 0.10 to 0.35 Number of Laskin Nozzles needed at 10 psig 1 1 1 Exhaust HEPA Filter w/ Gasket Dimensions (inches) 18x18x3.31 30x18x3.31 42x18x3.31 *Motor Horsepower (HP) 0.13 0.13 0.30 *Capacitor (MFD, Volts) 10MFD, 115V (2) 4MFD, 230V
15MFD, 115V 4MFD, 230V Fluorescent Lights F15T8-SP35 F17T8 F25T8




General References American Conference of Governmental Industrial Hygienists. Industrial Ventilation, A Manual of Recommended Practice, Cincinnati, OH ASHRAE Standard Committee. ASHRAE Standard Atlanta: ASHRAE Publications Sales Department, 1995 British Standards Institution, Laboratory Fume Cupboards. Parts 1, 2 and 3, London: 1990 Department of Labor, Occupational Safety and Health Administration, 29 CFR Part 1910, Occupational Exposures to Hazardous Chemicals in Laboratories, Final Rule. Vol. 55, No. 21. Washington D.C.:1990 DiBerardinis. L. et al. Guides for Laboratory Design, Health and Safety Considerations. Wiley & Sons, 1987 McDermott, Henry, Handbook of Ventilation for Contaminant Control, 2nd Edition. Butterworth Publishers, 1985. Miller, Brinton M. et al. Laboratory Safety: Principles and Practices. American Society for Microbiology, Washington, D.C.: 1986 NIH Guidelines for the Laboratory Use of Chemical Carcinogens. NIH Publication No. 81-2385. Rayburn, Stephen R. The Foundation of Laboratory Safety, A Guide for the Biomedical Laboratory. Springer-Verlag, New York: 1990 Sax, N. Irving and Lewis, JR., Richard J. Rapid Guide to Hazardous Chemicals in the Workplace. Van Nostrand Reinhold, 1987. Schilt, Alfred A. Perchloric Acid and Perchlorates. The G. Frederick Smith Chemical Company, Columbus, OH: 1979. Steere, Norman. CRC Handbook of Laboratory Safety, 2nd Edition. CRC Press, 1971. Collins, C.H., 1988. Laboratory Acquired Infections, History, Incidence, Causes and Prevention, 2nd ed., Butterworths, London.

Kruse, R.H., WH. Puckett and J. H. Richardson. 1991. Biological Safety Cabinetry, Clin. Microbiol. Rev. 4:207- 241. Miller, C.D., D.H.M. Groschel, J. H. Richardson, D. Vesley, J. R. Songer, R. D. Housewright and W. E. Barkley. 1986. Laboratory Safety, Principles and Practices, American Society for Microbiology, Washington, D.D. National Research Council (U.S.) Committee on Hazardous Biological Substances in the Laboratory. 1989. Biosafety in the Laboratory. Prudent Practices for Handling and Disposal of Infectious Materials. National Academy Press, Washington, D.C. Rayburn, S.R. 1990. The Foundations of Laboratory Safety, Springer-Verlag, New York. U. S. Department of Health, Education and Welfare. Formaldehyde Decontamination of Laminar Flow Biological Safety Cabinets, National Institutes of Health, Division of Safety, Bethesda, MD 20892. U. S. Department of Health, Education and Welfare. Effective use of a Laminar Flow Biological Safe Cabinet. National Audiovisual Center (GSA), Sales Branch, Catalog No., NAC 005133, Washington, D.C. U. S. Department of Health, Education and Welfare. 1975. Selecting a Biological Safety Cabinet. National Audiovisual Center (GSA), Sales Branch, Catalog No. NAC 000709, Washington, D.C. U. S. Department of Health and Human Services. 1999. Biosafety in Microbiological and Biomedical Laboratories, 4th ed. U.S. Department of Health and Human Services, Publication No. (CDC) 93-8395. U. S. Government Printing Office, Washington, D.C.