Manual Supplement © 2008-2009 Fluke Corporation . All rights reserv ed. Printed in the U.S .A. Manual Title: Impulse 6000D/7000DP Users Supplement Issue: 3 Print Date: August 2007 Issue Date: 3/09 Revision/Date: 1, 1/08 Page Count: 10 This supplement contai ns information nece ssary to en sure the accuracy of the above manual.
Impulse 6000D/7000DP Users Manual Supplement 4/08 1 Change #1, 46470, 46697, 47714, 48007, 49680 Replace pages 31 through 35, with the following: Defibrillator Analyzer Specifications Energy Output Measurement Compatible Defibrillator Wa veshapes ....
Manual Supplement Impulse 6000D/700 0DP Users 2 8/08 Accuracy ............................................................. ± 0.1 ms Voltage Range ................................................................. 20 to 5000 V Accuracy ...............
Impulse 6000D/7000DP Users Manual Supplement 4/08 3 Scope Output Autorange ........................................................... 2000:1, 400:1 and 80:1: depe ndant on the range Waveform Playback Output ...........................................
Manual Supplement Impulse 6000D/700 0DP Users 4 3/09 Ventricular Fibrillation ......................................... Coarse and fine Monomorphic Ventricular Tachycardia ............... 120 to 300 (by 5) BPM Polymorphic Ventricular Tachycardia .....
Impulse 6000D/7000DP Users Manual Supplement 8/08 5 Normal Sinus waves Lead # I II III V1 V2 V3 V4 V5 V6 Ref. Amp. 70 % 100 % 30 % 24 % 48 % 100 % 120 % 112 % 80 % Lead I reference Performance waves and R wave detection Lead # I II III V1 V2 V3 V4 V5 V6 Ref.
Manual Supplement Impulse 6000D/700 0DP Users 6 4/08 Sine waves ......................................................... 0.05, 0.5, 5, 10, 40, 50, 60, 10 0, 150, and 200 Hz Pulse ................................................................... 30 and 60 BPM, 60 ms pulse width R-Wave Detection Waveform .
Impulse 6000D/7000DP Users Manual Supplement 4/08 7 Accuracy ......................................................... ±(10 % of setting + 0.2 mV) Amplitude of Transvenous Pacer Pulse Simulation signals relative to amplitude setting (in percent) Lead II reference Lead # I II III V1 V2 V3 V4 V5 V6 Ref.
Manual Supplement Impulse 6000D/700 0DP Users 8 4/08 Change #2 On page 12, prior to Analyzing Pacemakers (7000DP only) , add the following section: SCOPE OUTPUT Each time a defibrillator is fired into.
Impulse 6000D/7000DP Users Manual Supplement 4/08 9 Pulse playback +4 V, 1 ms pulse Start of defibrillator pulse playback [1] [1] Pulse playback includes 0.4 ms before th e pulse was detected and theref ore displays any ear ly transitional signals. Use the synchronizing marker pulses to display specific information from the Scope Outpu t signal.
Manual Supplement Impulse 6000D/700 0DP Users 10 8/08 Change #3, 48038 On page 38, under Pacemaker Input, change: From: Accuracy …………..±1 % non-inductive (<2 μ H) To: Accuracy………….
An important point after buying a device Fluke 7000DP (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Fluke 7000DP yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Fluke 7000DP - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Fluke 7000DP you will learn all the available features of the product, as well as information on its operation. The information that you get Fluke 7000DP will certainly help you make a decision on the purchase.
If you already are a holder of Fluke 7000DP, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Fluke 7000DP.
However, one of the most important roles played by the user manual is to help in solving problems with Fluke 7000DP. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Fluke 7000DP along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center