Victoreen ® 05-437 Primalert ® 35 Area Radiation Monitor Operators Manual March 2005 Manual No. 126011 Rev. 3 © 2003, 2005 Fluke Corporat i on, All rights reserved .
Fluke Biomedical Radiation Management Services 6045 Cochran Road Cleveland, Ohio 44139 440.498.2564 www.flukebiomedical.com/rms.
i Table of Contents Section 1: Introductio n ................................................................................................ 1-1 1.1 Product Descr iption ................................................................................
(Blank page).
Introduction Product Description 1 1-1 Section 1 Introduction 1.1 Product Description The Primalert® 35 Area Ra diation Monitor provides reliable, continu ous monitoring of radioacti ve materials. Six front panel, color-coded LED's (1, 2, 4, 8,16, and 32 mR/h) indicate an increase or decrease in radiation level.
Victoreen 05-437 Operators Manual 1-2 1.2 Specifications Detector Energy -compensated G-M tube Energy Dependence 10% to 40% from 50 KeV to 2 MeV, as referenced to 137 Cs Overload Protection Alarms in fields higher than 100 R/h wh en detector is completely blocked fo r pulse operation.
Introduction Specifications 1 1-3 Input 230 VAC, 50 Hz, 13.8 W Output 12 VDC, 580 mA Part 14-417, Part 14-416 (Australia) Dimensions 5.9 in. high x 3.4 in.
Victoreen 05-437 Operators Manual (Blank page).
Getting Started Receiving Inspection 2 2-1 Section 2 Getting Started 2.1 Receiving Inspection Upon receipt of the unit: 1. Inspect the carton(s) and contents for d amage . If damage is evident, file a claim with the carrier and notify Fluke Biomedical, Ra di ation Management Services at 440.
Victoreen 05-437 Operators Manual 2-2 2.4 Optional Power The Primapak II storage battery sup plies a source of 12-16 Volt DC powe r to the Primalert using the converter plug conne ction. The plug is a Switchcraft #760, or e quivalent, with the sleeve negative and pin positive.
Operation Set Up 3 3-1 Section 3 Operation 3.1 Set Up The only set-up required for the Primalarm is the att a chment of one end of the power cord (converter) to the unit and the other end into a properly gr ounded AC outlet as noted in Section 2. When power is first applied to the unit, the alarm and one or more of the LED's may light.
Victoreen 05-437 Operators Manual (Blank page).
Maintenance Maintenance 4 4-1 Section 4 Maintenance 4.1 Maintenance This instrument contains CMOS integrated circuits. No service should ever be attempted unless by a qualified service technici an thoroughly familiar with these devices.
Victoreen 05-437 Operators Manual 4-2 Table 4-2. PC Board Assembly (126021) Parts List Item No. Ref. No. Part No. Description Qty 6 V1 010054 Tube Shield Assembly 1 7 U1, 4 630001 IC, Quad Nor Gate, 4.
Maintenance Maintenance 4 4-3 46 R22, 23 302041 Resistor, Film, 200k, .25W, 5% 2 47 R20, 26,30, 32,33,38, 341041 Resistor, Film, 100k, .25W, 5% 6 49 R27, 29,31, 34,40 341032 Resistor, Film, 10k, .25W, 5% 5 50 R13-18, 28 185-1448 Resistor, Film, 200, .
Victoreen 05-437 Operators Manual 4-4 Figure 4-1 Primalert 35 Assembly (126028).
Maintenance Maintenance 4 4-5 Figure 4-2 Primalert 35 Schematic (126015).
Victoreen 05-437 Operators Manual 4-6 Figure 4-3 PC Assembly (126021).
.
Fluke Biomedical Radiation Management Services 6045 Cochran Road Cleveland, Ohio 44139 440.498.2564 www.flukebiomedical.com/rms.
An important point after buying a device Fluke 05-437 (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Fluke 05-437 yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Fluke 05-437 - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Fluke 05-437 you will learn all the available features of the product, as well as information on its operation. The information that you get Fluke 05-437 will certainly help you make a decision on the purchase.
If you already are a holder of Fluke 05-437, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Fluke 05-437.
However, one of the most important roles played by the user manual is to help in solving problems with Fluke 05-437. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Fluke 05-437 along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center