When calling your representative for technical support, please have your serial numbers available. The Sensor and Instrument Serial Numbers are on the instrument, also see section 3.4.4.4.
Your Representative is:
Except as may be provided by contract, this document and all specifica- tions and drawings contained are the property of Xentaur Corporation, are issued in strict confidence, and shall not be reproduced or copied or trans- mitted, in any form or by any means, or used as the basis for the manufac- ture or sale of apparatus, programs, or services without permission.
Check the Internet for updates; the latest revision of this manual is available in Adobe Acrobat for- mat at: http://www.xentaur.com
Document No.: XDO.01.D.0000 Rev.2.1 5/10/02
Copyright © 2000 by Xentaur Corporation

Xentaur reserves the right to change or modify the product specification and / or appearance at any time without notice. Therefore, the information in this document is subject to change without notice and does not represent a commitment on the part of Xentaur Corporation.
The customer agrees that in accepting and using this instrument Xen- taur Corporation’s liability arising from or in any way connected with this instrument shall be limited exclusively to performing a new calibration or replacement or repair of the instrument or sensor, at Xentaur’s sole option, as covered by Xentaur’s warranty. In no event shall Xentaur be liable for any incidental, consequential or special damages of any kind or nature whatsoever, including but not limited to lost profits arising from or in any way connected with this instru- ment or items hereunder, whether alleged to arise from breach of contract, express or implied warranty, or in tort, including without limitation, negligence, failure to warn or strict liability.
Swagelok, Cajon are trademarks of SWAGELOK Co. Acrobat is a trademark of Adobe Systems Incorporated Epson is a registered trademark of Seiko Epson Corporation Microsoft Windows is a registered trademark of Microsoft Corporation HTF is a trademark of Xentaur Corporation SpanCheck is a trademark of Xentaur Corporation The Xentaur Logo is a trademark of Xentaur Corporation

Examine the XDT package for damage or mishandling. If any damage is evident notify the carrier and request an inspection.
Unpack the box, it should contain: The XDT, sensor in desiccant container, connec- torized cable, and this manual.
PLEASE READ THIS MANUAL IN WHOLE, PRIOR TO INSTALLING OR REMOVING THE SENSOR FROM ITS SHIPPING CONTAINER.
This manual is organized in three sections: Section 1 is an overview of the XDT. Section 2 describes the sensor and sampling techniques. Section 3 describes the instrument’s electrical, mechanical, and user interfaces.
This manual is intended for those already familiar with the installation, use and maintenance of analytical or process instrumentation.
Those acquainted with other Xentaur dewpoint measurement products such as the LPDT or the XPDM, will benefit from the commonality of the user interface.
Bornier de L’Ecran de Protection
Schutzerde
CAUTION - Risk of Electric Shock ATTENTION - Risque de Décharge Électrique
ACHTUNG - Hochspannung Lebensgefahr

Maintenance Policy In cases when equipment fault is suspected, please notify your representative of the problem, be sure to provide them with model and serial numbers. If the problem can not be resolved, then ask for a Return Authorization Number (RAN) and shipping instructions. Issuance of an RAN does not automatically imply that the equipment is covered by our warranty, that will be determined after we receive the equipment. Pack the equipment in a suitable box with sufficient padding, include the RAN number on your paperwork, and send the equipment, prepaid, to the designated address. Equipment returned without an RAN, or with reversed shipping or import/export charges, will not be accepted If the warranty has expired, or the damage is due to improper use or exposure of the equipment; then the repair facility will provide an estimate and wait for approval before commencing repairs.
For your convenience a Return Authorization Request Form is provided in appendix N, it must be com- pleted and sent back to the provided destination in order to obtain a RAN.



The XDT is a microprocessor based hygrometer, for measuring moisture content in gases in the range from -100°C to +20°C dewpoint depending on the sensor ordered. The XDT uses the Xen- taur Hyper Thin Film HTF™ sensor which is encapsulated in sintered stainless steel, thus it is capable of coming into contact with a wide variety of environments. However one should keep in mind that the sensor is a delicate device and it should be handled accordingly. The measurement is displayed on the instrument’s custom LCD, can be transmitted by optional analog and digital outputs, and can control optional programmable relays. Four front panel but- tons provide the user with a rich feature set. The XDT’s advanced design allows it to be housed in a variety of enclosures. The XDT has an impressive set of dewpoint measurement capabilities in terms of accuracy, stabil- ity, response time etc. The specifications of the sensor are discussed in section 2.2, while the spec- ifications of the instrument are discussed in section 3.2. A summary of the standard and optional features & capabilities of the XDT are listed below as an overview aid to the user.



2.1 Precautions using the sensor

Type: Hyper Thin Film HTF™ high capacitance Al2O3. Dewpoint range: XTR-100: -148°F to +68°F (-100°C to +20°C) XTR-65: -85°F to +68°F (-65°C to +20°C). Capacitance: 15nF to 200nF. Accuracy: ±5.5°F (±3°C) refer to appendix M for accuracy expressed in other units of measure. Repeatability: ±0.9°F (±0.5°C). Response time: refer to Dewpoint Response time analysis in Appendix G. Operating Temperature: 14°F to 158°F (-10°C to +70°C), non-condensing Storage Temperature: -40°F to+176°F (-40°C to +80°C), non-condensing Sample Flow range: (linear velocity @ 1ATM): Static to 100m/s. Enclosure: encapsulated in 100µ sintered stainless steel. Calibration method: Highly uniform sensors calibrated at low dewpoint and SpanCheck™, sensor saturates at dewpoint
above +68°F (+20°C). NIST/NPL traceable multi-point factory calibration available optionally. Pressure operating range:...Standard:500 PSI (34 bar).
Optional:5,000 PSI (340 bar). Mechanical connections: 14mm x 1.25mm sparkplug threads, and ¾”-16 threads, standard
Optional configurations: G1/2, 1/2”NPT, 5/8”-18 and others Electrical connections: Female BNC connector. Sensor signal cable: RG58 coaxial cable, or for lengths greater than 100’ RG6 coaxial cable, max 3,000’. Approvals/Classifications:.CE for electromagnetic compatibility, accredited laboratory tested and certified
UL and cUL for ordinary use Intrinsically Safe configurations: for NEC standard IS Simple Apparatus: UL for CENELEC standard IS Simple Apparatus: DEMKO, refer to labeling
Keep in mind that the moisture content at the sensor is not only due to the moisture of the gas being measured, but also due to desorption of water from tubing, trapped moisture (at the inter- connection points, valves, filters and other hygroscopic materials in the system), leaks in the sys- tem, and others. Thus the measurement may vary from the expectation, and therefore care should be taken in choosing the sampling technique utilized in the measurement. Factors such as gas pressure, flow rate, materials of construction, length and diameter of tubing, number of intercon- necting fittings, dead space in tubing and manifolds; will influence the measurement value and response time. The high capacitance HTF™ sensors can be installed either directly in the line to be sampled (in- situ), or in a slip stream of a sample system (extractive). To assure a long and accurate performance of the sensor, it should be protected from contaminants such as liquids (water, oil etc.), and particulates. The sintered stainless steel sensor encapsulation protects from particulates larger than 100 microns, finer particulates (e.g. from degraded desic- cant or rust) should be filtered with a particulate filter with suitable capability, do not use hygro- scopic filter materials. Refer to Sample Gas Filter Considerations Appendix H.


* maintain differential pressure to provide adequate flow through sample cell
The sensor measures the dewpoint at whatever pressure it is at. The instrument is equipped with a pressure correction feature which allows reporting the dewpoint at a different pressure.

2.3.2 Extractive Installation For extractive installations we recommend our sample system ESS, which may be equipped with a variety of features, such as: isolation valve, coalescing or particulate filter, pressure regulator, calibration sample injection or extraction port, pressure gauge, flow meter, weatherproof enclo- sure. Refer to the ESS literature for more information. If the resources to make your own sample system are available, the following two diagrams may be used as a guideline to configure a simple system.


The sensor probe has two thread sizes for mounting to the sample cavity where the dewpoint will be measured, see appendix C. Various optional fittings are available for direct connection into existing system openings, refer to appendix D. Ask your representative for a Sample Cell, if you do not have the ability to provide an appropriate sample cavity mounting. If the ¾”x16 thread is used then the sensor will seal against the wall of the sample cell with the provided Viton A O-ring. If the 14mm x 1.25 spark plug thread is used then an additional Viton gasket must be installed to provide the proper seal. Ask your representative for this gasket, it is available free of charge. To prevent any leaks, tighten the fitting into the sample cavity, with a 11/4” wrench, 1/8 turn past fin- ger-tight to assure metal-to-metal contact. The sensor can be removed from the fitting by unscrewing it. Make sure that the sensor is securely fastened to the fitting (the tension washer should be compressed), so that it does not come loose during use.

If erroneous readings are suspected on a newly acquired instrument, compare the serial number engraved on the sensor sintered filter, to the label on the instrument. The two should be the same; if they are not, the instrument may not be calibrated with the installed sensor. To troubleshoot other problems, identify the unexpected reading category in the following table, and consider the possible causes and appropriate diagnostic action and remedy.
Symptom Possible Cause Diagnostic/Remedy
Condensation will occur if the temperature of the sample system, at any point is below (colder) the dewpoint temperature of the sample gas. Once having formed, the sample reaching the sensor will have a dewpoint equal to the temperature of the condensation, regardless of the dewpoint of the source gas.
Condensation in sample system.
It is usually more satisfactory to bleed a sample gas at atmospheric pressure through the sensor sampling chamber, and to use 1/8” (3mm) o.d. sample pipe.
See below re sample pipe material, also see section 2.3
Dry Reading SpanCheck™, wrongly set, or faulty Sensor. Verify SpanCheck™, or return sensor for full calibration to your representative.
Leak in system or use of unsuitable pipe. Cure the leak, or replace unsuitable pipe with copper or stainless steel. Flexible con- nections should be made with PTFE pipe. NEVER use rubber or plastic pipe.
Comparison of readings with manual cooled- mirror instrument.
This type of indicator reads about 10ºC dry at about -50ºC dewpoint due to tempera- ture gradients within the device. The error increases at drier levels.
Prolonged exposure to wet gas. Dry down the sensor, install sensor in either a known dry gas stream i.e. instrument quality air or dry nitrogen, or place sensor in a dry can or bottle of desiccant and seal the container from outside air (the shipping container is designed for this purpose) also see section 3.4.4.3.1
2.Short circuit on sensor cable or connections. Disconnect cable from sensor and if meter still reads SHR, cure the short circuit in the cable or connections or replace cable; otherwise check the sensor.
3.Short circuited sensor. Disconnect cable from sensor and note that the meter reading returns to OPN. Use a new sensor, or apply approximately 20V DC to the sensor MOMENTARILY with the sensor in a known dry condition. Polarity is not important, but the contact MUST be very brief or the sensor may be damaged. Consult with your representative for a cir- cuit diagram.
OPN. 2. Open circuit on cable. Disconnect cable from sensor and short center pin of plug to the outer shell. If the dis- play still shows OPN, repair cable.
For non-sensor related problems (e.g. no reading on instrument) refer to section 3.6

3.1 Precautions using the XDT

XDT-NEMA tabletop or surface mount, watertight IP65 tested, and NEMA 12 tested XDT-PM panel-mount, DIN 43700 dimensional standard, optional NEMA 12 type protection gasketing available Dimensions & Weight: XDT-OEM: 4.25” x 5.15” x 2.25”, 0.6 lbs (fully optioned) without cables.
XDT-NEMA: 4.73” x 6.3” x 3.55”, 2.1 lbs (fully optioned) w/cables. XDT-PM: 5.67” x 2.84” x 2.95” DIN 43700 standard, 1.3 lbs (fully optioned) w/cables. refer to appendix E for detailed dimensions Environmental Range: Operating temperature of electronics: 14°F to 122°F (-10°C to 50°C)
Storage temperature of electronics: -40°F to 176°F (-40°C to 80°C) Humidity: 0 to 90% RH non-condensing Altitude: 0 to 6500 feet (2000 meters) Mains Supply Voltage: 100 to 250VAC autoranging, 50/60Hz, 10VA, internal 0.5A, 250V fast acting fuse.
Optionally the instrument may be configured with DC power capability: 15 to 30VDC 0.5A. Electrical connections: 21 contact pluggable screw terminal block, 1.3mm diameter pins on 5mm centers. Wiring requirements: For AC Power: 18AWG or heavier wire, an external means for disconnecting the power source is
required to meet National Electrical Code requirements. For Sensor signal: RG58 coaxial cable, or RG6 for cable longer than 100’. A 6 foot cable is provided as a standard, other lengths ordered as options. When changing cable, refer to cable compensation section. Input resolution: 0.1°C dewpoint. Indicators: 3.5 digit backlit LCD with custom legends, audible indicator. Engineering units: °C,°F, PPM, LBS H2O/mm scf,gm H2O/m3. Controls: Four push buttons, user’s selections are stored in EEPROM. Outputs: Analog: voltage by order or current user selectable 4-20mA or 0/24mA. Linear to the selected engi-
neering units, the range is programmable. Output resolution is 0.1°C dewpoint, linearity 1%, max load resistance 500 Ohms. Digital RS-232C (9600,8,E,1), can interface to a Personal Computer or other RS-232 device. Alarm relay contacts: Ordinary use, explosion-proof housed, and safe area instruments: 10A, 250VAC or 30VDC.
Intrinsically Safe Div 2 instruments use hermetically sealed relays: 3A, 120VAC Isolation: Sensor is isolated from power ground, analog output and RS-232, but they are shunted with a 33V tran-
sorb, a 1M Ohm resistor and 0.1uF capacitor. Refer to Electrical Connections section. Approvals/Classifications: CE for electromagnetic compatibility, accredited laboratory tested and certified
UL and cUL for ordinary field wired use Intrinsically Safe configurations: for NEC standard IS installations UL, for CENELEC standard IS installations DEMKO, refer to instrument labeling

3.3.1 Instrument Mechanical Installation The XDT is available in several different physical configurations. Please follow the instructions below, which describe the instrument being installed. The sensor installation is discussed in Sec- tion 2: Sensor and Sampling Techniques.
3.3.1.1 XDT-PM (DIN43700) Enclosure Installation The standard XDT-PM is provided with two clip-on compression panel mounting brackets. These brackets allow the user to mount and secure the instrument onto a properly cutout panel (137mm x 67mm). Consult with drawing in appendix E, for all relevant dimensions. Installation is accom- plished by attaching 2 clips, one on each side of instrument, and inserting the instrument into the panel cutout. Pressure must be applied along edges of instrument until it is firmly seated.
3.3.1.2 XDT-NEMA (IP65) Enclosure Installation The instrument can be installed as a wall or panel surface mount, by making use of four screw holes on the back (bottom) of the instrument enclosure. Open the instrument cover for access to these screw holes. They are located at the corners of the enclosure and outside of the NEMA seal. Do not drill other mounting holes, as you may compromise the seal. Refer to Appendix E for all relevant dimensions.
3.3.1.3 XDT-OEM bare board Installation When mounting the circuit board in a user supplied enclosure using standoffs, be sure to use insu- lating washers. Refer to Appendix E for mounting hole dimensions and required mounting dimen- sions.

3.3.2 Electrical Connections All connections are made via a 21 contact, pluggable screw terminal block referred to as P1.
2 AC NEUTRAL 1 AC LIVE
275VAC Power Supply Module
4.3mH
0.1uf
0.0047uf 0.0047uf
5 SIG-IN 6 SIG-RET
SENSOR CABLE
0.1uf
1M Ohm 33V 400W
11 RET 10 RxD 9 TxD 8 A-RET 7 A-OUT
Logic ground and AC Power
ground are connected for safety
13 AC GROUND
17 C 16 NO 15 NC
LO
21 C 20 NO
HI
Note: * Options will function only if they are installed
3.3.2.1 Connecting Power The XDT is typically provided to be field wired to operate from AC Mains Power of 100 to 250VAC, 50/60Hz, 10VA, the instrument has an internal 0.5A, 250V fast acting fuse. However the XDT can also be optionally ordered to operate from 15 to 30 VDC 0.5A, with an internal 0.5A

fast acting fuse. Please follow the appropriate installation procedures in the paragraphs below according to the electrical power option on your instrument. Do not connect to AC power without a proper ground connection. For Intrinsically Safe installations, first make sure that the equipment is qualified for the particu- lar installation, it should have a label specifying the certifications and the approving agency. Then follow the appropriate control drawing as well as the instructions in the relevant paragraphs in this manual.
3.3.2.1.1 AC Mains Electrical Power Connection The XDT is provided with a 6 foot (2 meter) internationally approved AC power cord, terminated with a connector according to the ordered option. This cable is connected to the pluggable screw terminal block; however it is provided strictly as a means of testing the instrument, it should not be used in the final installation. The XDT is intended as a field wired instrument permanently connected and installed according to the local, nationally recognized procedures for equipment of this type and stated power requirements. For North American installations use UL 62 and/or CSA C22.2 No 49, 18AWG cable with black, white, and green color codes. For European and other international installations use CENELEC harmonized type cable, with 0.82mm wire size or equivalent 10 amp use, with brown, light blue, and green/yellow stripe color codes. The power wiring must be connected to the pluggable screw terminals marked AC LIVE (#1), AC NEU- TRAL (#2), and AC GROUND (#4). A switch or circuit breaker shall be included in the installa- tion. It shall be in close proximity to the equipment and within easy reach of the operator. It shall be marked as the disconnecting device for the instrument. For the XDT-NEMA use the 1/2” NPT conduit fitting as the power cable access, do not use the cable grips they are for low voltage use only.
3.3.2.1.2 Low Voltage DC Powered Option - Electrical Power Connection When an XDT is configured with the low voltage DC power option, it can be powered by apply- ing 15 to 30 VDC 0.5A, to the PS+ (#12) and RET (#11) pluggable screw terminals. The positive supply must be connected to PS+, while the negative or ground to RET. The AC terminals must not be connected. Note that the DC supply is ground referenced to the analog output and RS-232 grounds.
3.3.2.2 Sensor Connection The sensor input terminals are labeled “SIG-IN” (#5) for the center core of the coaxial sensor cable, and “SIG RET” (#6) for the outside braid of the cable. Factory supplied coaxial cables have a BNC connector at one end to mate to the sensor, and pigtails at the other end to be placed in the screw terminals of P1, the black pigtail is the braid and should be connected to the terminal marked “SIG RET”. The coaxial cable can be as much as 3,000 ft. long, however if the cable is changed for a longer or shorter one the instrument must be compensated for the new cable (see section 3.4.5 -11). The instrument is properly compensated for the cable supplied from the fac- tory. Consult your representative for obtaining the proper cable.
3.3.2.3 Wiring the Alarm Contacts The optional alarm relay contacts are located on the terminal strip P1. The terminal strips are marked to indicate wipers and normally open and normally closed contacts, of the two indepen-

dent relays corresponding to the HI and LO alarms. The relay contacts are rated at 10A 250VAC or 30VDC, instruments for use in division 2 Hazardous Areas have hermetically sealed relays rated at 3A 125VAC. Use wiring appropriate for the voltage and current that will be switched by the relays. Also see section 1.3 for EMI considerations. Keep in mind that the relay polarity is programmable thus wiring should be designed to provide a fail safe operation in case of power failure. See section 3.4.4.2. Also note that while viewing the dewpoint, the display will flash HI and/or LO as necessary to indicate that the corresponding alarm relay is de-energized. Specially ordered instruments may have a third alarm, refer to the supplied addendum for relay contact rat- ing, pinouts and user interface issues.
3.3.2.4 Interfacing to the Analog Output The optional Analog Output may be a voltage or current, depending on the ordered configuration. The Analog Output signal is provided on the P1 connector, when wiring please observe the polar- ity indications. The positive terminal is on pin #7 it is labeled “A-OUT” and the negative is on pin #8 it is labeled “A-RET”. The ground of the Analog Output is connected to the frame (AC power) ground but it is isolated from the sensor ground, see section 3.1.1. Instruments ordered with Voltage output, are configured at the factory with appropriate resistors across the current loop output such that when operated in the 0-24mA mode they produce the desired voltage. The current loop is capable of driving loads from 0Ω to 500Ω and the user may select to operate it as 0-24mA or as 4-20mA (refer to 3.4.5-6). The output is linearly proportional to the selected engineering units. The output may be scaled such that it spans only a portion of the full range of the sensor, this feature may be useful in cases where a higher resolution output is required over a narrow dewpoint range, or vise versa. To verify or change the current loop configuration and scal- ing follow the instructions in the set-up mode section 3.4.5 - 6,7,8&9. After hooking up the current loop output, it can be forced to its low, mid and high points by fol- lowing the instructions in section 3.4.5 - 7. This procedure may be helpful in testing the connec- tion and setting-up the termination equipment.

3.3.2.5 Interfacing to the RS-232 option The optional RS-232C interface is provided on the P1 connector. The configuration is 9600 baud, Even Parity, 8 Bits, 1 Stop, all received characters are echoed. The ground of the RS-232C inter- face is connected to the frame (AC power) ground however it is isolated from the sensor ground, see section 3.1.1. To connect the instrument to a Personal Computer a 3 conductor cable is required; with wires to be placed in the screw terminals of the XDT at one end, and with either a DB9 or DB25 female connector at the other end:
Signal Name instrument P1
pin # DB9 pin or DB25 pin
Transmit Data 9 TXD or Tx 2 3
Receive Data 10 RXD or Rx 3 2
Signal Ground 11 GND 5 7
RTS n.c. 7 4
CTS n.c. 8 5
DTR n.c. 4 20
DSR n.c. 6 6
Note that some Personal Computer Programs may require that RTS & CTS and/or DTR & DSR are jumpered for proper operation. This jumpering may be accomplished at the DB9 or DB25 connector. CAUTION: Connecting the RS-232C interface on instruments which do not have the option installed, will damage the instrument. Refer to appendix K for details on the protocol used on the RS-232C interface.
3.4.1 Starting up The instrument is ready for use as soon as the power cabling is connected. When power is applied the instrument will initialize its program and for a moment display XEN, then it will enter the Operating State. If the MODE button is held pressed while the instrument is performing its power-up initialization, it will enter the Set-Up State, which allows the user to select setup vari- ables of the instrument.

3.4.2 Display Conventions 1. To display characters with the 7 segment numeric display, the following pseudo-alphanumer- ics are used:
Numbers: 0 12 3 4 5 6 7 8 9 0 1 2 3 4 5 6 7 8 9 Letters: ABCDEFGH IJ LNOPQRS TUXYZ A B C D E F G H I J L N O P Q R S T U X Y Z Symbols: ? - .? - . 2. The instrument will indicate whether a particular mode allows changing a parameter by show- ing the word “SET” in the upper left corner of the display. Be careful not to change any parameter inadvertently. 3. °C°F appear simultaneously, to indicate the sensors’ attenuation in decibels. 4. Values larger than ±1999 or smaller than ±0.01 are displayed in powers of 10±3. As required, either a “10 3” or “10-3” will appear above and to the right of the displayed value, the value will be rounded off to 3½ digits. The display will show RNG (out of ranGe), if the number to be displayed is larger than 1,999,000. Twelve examples follow; for each the number and units desired to be displayed are shown in italics, depicted immediately below them is the resultant 3½ digit LCD display:

3.4.3 Push Buttons Four push buttons provide user control of the instrument. They are designated MODE, UP, DOWN and PRESSURE CORRECT. These push buttons are vertically lined up on the upper left side of the bare circuit board, with the MODE button on top. There is also a connector provided which allows the use of external switches. The XDT-NEMA makes use of the circuit board mounted push-button switches, to access them one must open the cover of the XDT-NEMA enclosure. The XDT-PM is available in several versions: • With a solid silk screened front face plate, which must be removed to access the circuit board


Remove front plate to access control buttons, consult the manual.
Panel Mounted Instrument with a solid silkscreened face plate
Use probe (paper clip) to push control buttons, consult the manual.
Panel Mounted Instrument with a silkscreened face plate with holes
In general the MODE button navigates through the different user options “Modes”; the UP and DOWN buttons modify the units, values or choices in the selected mode. Refer to the flow dia- grams in Appendix A for detailed overview of button functionality. A button may be held down for a prolonged time, for accelerated incrementing or decrementing of numeric values.
3.4.4 Operating State Upon power up, the unit performs certain initialization tests (see table in section 3.6), and enters the ‘Operating State’, in the Viewing Dewpoint mode. Depressing the ‘MODE’ button will change modes (see appendix A) in the following order: nViewing Dewpoint oAlarms pStart Calibration qViewing Serial Number (back to) nViewing Dewpoint. The unit will return to Viewing Dewpoint mode if no buttons are pressed for 30 seconds, unless it is performing a cali- bration.
3.4.4.1 Viewing Dewpoint Mode In this mode the user can view the dewpoint, this is indicated by the presence of the ‘DEW- POINT’ legend on the lower left of the display. The available engineering units in which to view the moisture content are °C, °F, PPM, LBS and G/M3; the UP and DOWN buttons scroll back and forth through these units in respective order. The °C and °F are dewpoint readings. The PPM is

parts per million by volume computed at the sensor pressure (more about pressure later). The LBS and G/M3 are density measurements, pounds of water per million standard cubic feet and grams of water per standard cubic meters, respectively, both in Natural Gas, they are computed accord- ing to data derived by IGT Research Bulletin 8, taking into account sensor pressure. Note that the analog output is linear to the selected engineering units, therefore be mindful that while scrolling through various units the analog output may change even though the measured dewpoint is stable.
The PSI legend flashes at the bottom of the display, when there is pressure correction in the com- putation of the displayed values. A short press of the ‘pressure correct’ button toggles the unit in and out of pressure correct mode. When there is no pressure correction applied, the PSI legend does not appear. Sensor Pressure is used in the context that this is the pressure inside the sample cell when per- forming the measurement, i.e. it is the operating pressure of the sensor. Gas Pressure is used in the context that this is the pressure at which the dewpoint is to be calcu- lated. Pressure Correction is used in the context that the values displayed signify the moisture content at some pressure (we refer to this as the ‘Gas Pressure’) different from the pressure at the sensor. Note that PPM, LBS and G/M3 readings are by definition unaffected by pressure correction because only the pressure at the sensor affects their value. While °C and °F are affected by pres- sure correction by reporting what the dewpoint would be at the Gas Pressure when the dewpoint is what is measured at the pressure at the sensor. However, this also implies that whether pressure correction is applied or not the PPM, LBS and G/M3 readings are affected by the setting of the sensor pressure. A long press of the Pressure Correct button, while in the pressure correct mode (flashing PSI leg- end), changes the unit to the View/Set Sensor Pressure sub-state. The display has the ‘SET’ and ‘PSI’ legends on, and alternately shows SEN and the currently set value for the sensor pressure. The up and down buttons allow the user to modify the sensor pressure, while a short press of the pressure correct button toggles the Sensor Pressure setting between whatever value is on the dis- play and 14.7 psi this is a quick way to go back to atmospheric settings. A long press of the pressure correct button changes the unit back to the Viewing Dewpoint Mode. Pressing the ‘Mode’ button changes the unit to the View/Set Gas Pressure sub-state. The display has the ‘SET’ and ‘PSI’ legends on, and alternately shows GAS and the currently set value for the gas pressure. The up, down and pressure correct buttons operate in the same manner as in the Sensor Pressure sub-state. Pressing the ‘Mode’ button changes the unit back to View/Set Sensor Pressure sub- state, and so forth. Notes: 1. Instruments are shipped from the factory in the locked mode and must be unlocked before this procedure can take place (see section 4.3 #7 to unlock). 2. The factory default settings are: 14.7psi for both sensor and gas pressure and pressure correction disabled. 3. When Pressure correction is disabled all dewpoints are computed by assuming that both Sensor and Gas Pressures are 14.7psi. 4. Pressure correction will effect the analog output but not the alarm relay trip-points. The analog output will track the pressure corrected dewpoint, the relays will trip on the uncorrected dewpoints.

3.4.4.2 Alarms There are two independent optional alarms, they are named HI and LO alarms. Each alarm can activate a single pole double throw relay rated at 10A 250VAC or 30VDC per contact. Instru- ments for use in Division 2 Hazardous Areas have hermetically sealed relays rated at 3A 125VAC. Specially ordered instruments may have a third alarm, refer to the supplied addendum for relay contact rating, pinouts and user interface issues. The alarms can be set with a trip-point at any dewpoint within the range of the selected sensor. There is also a selectable hysteresis (with a minimum value of ±0.5°C to prevent relay chatter) which allows driving systems such as regen- erative dryer purge valves in “dewpoint demand mode”. The polarity of the alarm is also select- able, thus one may choose whether the relay energizes above or below the trip point, to allow fail safe design in case of XDT power loss, or in case of any other errors or failures which will cause the relays to de-energize. When an alarm relay is deenergized the corresponding HI and/or LO indicator flashes on the dis- play while viewing the dewpoint.
The behavior of the alarm when a sensor failure (e.g. open or short) is detected is also program- mable. The options upon sensor failure are: 1. Fail High - put the alarm in a state as if the dewpoint is high, e.g. A.X.H 2. Fail Low - put the alarm in a state as if the dewpoint is low, e.g. A.X.L 3. Fail Flashing - Energize/Deenergize the relay alternating once every 2 seconds, e.g. A.X.F 4. No Special Handling - if sensor is open the alarm is in a low dewpoint state; if the sensor is shorted, the alarm is in a high dewpoint state, e.g. A.X.N
Setting or checking the present setup of the alarms is illustrated with the following example: EXAMPLE : - Set the ‘HI’ alarm to de-energize the relay when the dewpoint is wetter than -75°C with minimal hysteresis (±0.5°C), and faults such as sensor failure cause relays to react as if there is high dewpoint i.e. if sensor cable breaks the relay de-energizes. When following these instructions, it may be helpful to refer to Appendix A.
the asterisks ‘*’ take the place of characters that may appear depending on previous set- tings; the ‘SET’ indicator means that changes can be made; the ‘HI’ indicator means that we are changing the HI Alarm, (alarm #1). Note that if the instrument does not have the alarm options installed this MODE (User Option) will not appear.
the -B means energize below or de-energize above set dewpoint, the H means faults look like high dewpoint. If the display shows LOC, the alarm changes are locked out, repeat step 1.

the ‘DEWPOINT’ indicator means that we are changing the alarm trigger dewpoint
˚C DEWPOINT
the Z indicates that we are changing the hysteresis of the alarm.
˚C
The Alarm Relays switch on non-pressure corrected dewpoints. Thus enabling or disabling the pressure correction wil not effect the relays.
The instrument will retain the alarm settings even if the power is turned off.
Pressing the MODE button changes to the Start Calibration Mode.
In general we recommend to perform SpanCheck in 12 month intervals (not more often), and sin- gle point calibration as often as practically possible. When it is time to perform both calibrations, then first disable the single point calibration, then perform the SpanCheck calibration, then per- form the single point calibration. This procedure is not necessary but it is preferred when both cal- ibration are performed.

3.4.4.3.1 SpanCheck™ Mode The instrument is calibrated at the factory with the sensor it is shipped with and does not need to be re-calibrated prior to installation.
Instrument calibration is recommended in approximately 12 month intervals, and the XTR65W sensor should be re-calibrated after prolonged exposure to liquid water. Simply follow steps 1 - 7 of the procedure below, removing the sensor from the sample gas stream. To avoid injury, make sure the gas stream is depressurized before removing the sensor. It is recommended to keep the sensor exposure to room air as short as possible, in order to avoid super saturation of the sensor. While super saturation is not damaging to the sensor, it will prolong the initial dry-down time after you install the sensor in the sample stream. Therefore, remove the sensor from the packaging container only after you are ready to proceed with the calibration pro- cedure and install the sensor in the sample stream immediately after the calibration procedure is completed. If you are not ready to use the sensor right away after calibration, put the sensor back in the shipping container for dry storage. The instrument must be calibrated with the sensor it will be used with.The calibration procedure takes advantage of the sensor’s ability to saturate (refer to Appendix F) and is executed by the instrument computer, by performing the following steps: If a single point calibration is in effect with a reference dewpoint below -70°C, it should be dis- abled (see next section) before performing SpanCheck calibration. 1. If the instrument is locked and a calibration is attempted; it will display LOC and will not per- form the calibration. To unlock the instrument consult section 3.4.5.-10 2. Push the MODE key a few times until the display shows CAL. 3. Press the UP button. The display will show CNF, prompting you to confirm that you want to start the calibration procedure. You can abort the calibration procedure by pressing the MODE key. 4. Remove the sensor from its packaging container or sample stream, so that the porous metal filter is visible and the sensor is exposed to in-hand micro-climate (refer to explanation below). Close the packaging container as soon as you have removed the sensor to avoid degra- dation of the desiccant inside the container. You may want to re-use the container at a later date. 5. Expose the sensor to in-hand micro-climate, and push the UP button again to confirm that you want to start the calibration procedure. The display will flash AC for 60 seconds, while the sen- sor is saturating. Make sure you keep the sensor exposed to in-hand micro-climate until the display shows END. 6. After 60 seconds, the display will flash the selected sensor type (see Selecting a sensor type 3.4.5.-2) and then the instrument will calculate the slope and offset of the sensor curve while displaying the calculations and then will display END for a few seconds, after which the instru- ment will automatically go into measuring mode. The sensor calibration is completed. (The display may flash SAT, indicating that the sensor has super-saturated. As soon as the sensor is exposed to an atmosphere with a dewpoint lower than the saturation dewpoint, the display will indicate the dewpoint measured by the sensor.) The display may show alternating SEN and 2LO as an indication that the measured capacitance is too low to be from a saturated sensor, in this case make sure that the sensor is properly plugged in, and repeat the calibration procedure. 7. Install the sensor in the sample cell or adapter fitting or put it back into the packaging con- tainer for later use.

The instrument will retain the calibration even if the power is turned off.
Under certain conditions, an over (super) saturated sensor may need to be completely dried out before a calibration is performed. Symptoms of these conditions are a sensor that will not go through the SpanCheck™ function to the END display, or a sensor that will not dry down after cal- ibration. To dry, install sensor in either a known dry gas stream i.e. instrument quality air or dry nitrogen, or place sensor in a dry can or bottle of desiccant and seal the container from outside air (the shipping container is designed for this purpose). After a minimum dry out period of 24 hours, proceed with the calibration procedure of your choice. SpanCheck™ calibrates sensors by using micro climates. Please contact your representative if you have any questions about how to create such micro climates. A perfectly acceptable and accurate for calibration micro-climate may be created for the XTR100 and XTR65 sensors by cupping the sensor in the palm of one’s hand during the SpanCheck™ pro- cedure. Remember that the micro-climate does not have to be accurate, it just has to be higher than the designed saturation level. The dewpoint of the micro-climate within the fingers is usually higher than the dewpoint of the room air and probably well above the +20°C upper range of the sensor. On a dry day, one may need to exhale in the hand before cupping the sensor to guarantee a high dewpoint. The sensor should be warmed up close to body temperature before performing this procedure. Care must be taken that the sensor temperature is not below the temperature of the hand, as in such case condensation could occur and super saturate the sensor. For the same reason, avoid exhaling directly onto the sensor.
3.4.4.3.2 Single Point Self Calibration, manual or scheduled The instrument is calibrated at the factory with the sensor it is shipped with and does not need to be re-calibrated prior to installation.
Single point calibrations should be performed whenever the measurement is in question, or on a scheduled basis depending on the application. Consult with your representative to establish crite- ria for specific applications. The single point calibration is performed on the premise that the sensor is exposed to a gas with a known dewpoint for a time sufficient to equilibrate. Then the sensor output is measured, and the calibration curve is adjusted to provide measurement computations which correspond to the known dewpoint. Measurements over the full sensor range are linearly adjusted to this known point. The single point calibration provided by this instrument may be performed manually, or (when a sample system is ordered equipped with an electrically actuated switchover valve) unattended on a repetitive timer scheduled basis. The user interface for both is the same, with additional steps for interval setting and automatic stabilization time outs for the instrument with the electrically actu- ated valve.



3.4.4.4 Viewing Serial Number Mode In this mode the user can view the instrument serial number. The display shows the serial number by alternately displaying XSN and the number. If the number is larger than 1999 then it is dis- played in 2 parts, first part is the thousands signified by the x103 legend in the upper right corner of the display and the second part is the units. For example serial number 12345 will be shown as:
Pressing the UP and DOWN buttons simultaneously resets the instrument, this is useful in instal- lations where power can not be turned on and off easily. Pressing the Pressure Correct button will cause the instrument to enter the setup state.
Pressing the MODE button, returns the unit to the Viewing Dewpoint Mode.
3.4.5 SetUp State To enter the Setup State power-up the unit (or reset it from the serial number Mode), and hold the MODE key pressed while the unit is initializing. Refer to Appendix B for a flow diagram of the SetUp State. The setup state provides eleven capabilities, each one controlled from its own mode. Depressing the MODE button navigates through the modes in the following order: nSelect Alternate Display Units oSelect Sensor Type pSet Measured Attenuation of Sensor at a Low Dewpoint qSet the Low Dewpoint at which the Attenuation was measured rSet the Calibration Mid-range Adjust sAnalog Output select 4/20mA or 0/24mA tPerform a Test and Verification of the Analog Out- put uSet dewpoint corresponding to low end of Analog Output vSet dewpoint corresponding to high end of Analog Output wLock/Unlock the instrument nnSensor Cable Compensation (back to) n. These eleven functions are explained in their respective sections which follow. Note that modes 6,7,8&9 will appear only on instruments with the analog output option. Changes can be made only to an unlocked instrument, the changes are stored (when the MODE button is pressed) and retained even if the instrument power is turned off. 1. Display of alternate units: In this mode, a second engineering unit can be chosen to be dis- played alternating with the engineering unit selected in the operating state, for example, a dewpoint can be alternately shown in °C and in PPM.

CAUTION: 2), 3), 4) and 5) are set at the factory and need only be modified when a sensor is changed. 2. Selecting the sensor type: In this mode the user can select the software matching the type of sensor installed in the instrument: XT1 : XTR-100 (-100°C to +20°C); XT6 : XTR-65 (-65°C to +20°C); DB : show impedance of any sensor, disable Open and Short error massages. 3&4Adjusting low end sensor attenuation and dewpoint: These modes are used to enter a data pair representing a low dewpoint and the sensor attenuation measured at this low dewpoint. (see Sensor Theory of Operation appendix F). This data pair should not be modified unless the sensor is replaced (refer to procedures for replacing a sensor in appendix L). The sensor ship- ping container is labeled with the proper values. 5. Calibration Adjustment: In this mode the user can enter a sensor specific Adjustment Value to improve the instruments’ accuracy in the range of -50°C to -10°C, this adjustment has very little or no effect outside this range. The Calibration Adjustment Value is printed on the ship- ping container supplied with the sensor, or may be derived if the sensor can be exposed to a known reference. Deriving an Adjustment Value: If the adjustment value for your sensor is not known, contact your representative and follow the instructions below for entering known values. If the value is not available, then it may be derived using a manual method, provided that a moisture cali- bration facility is available. The calibration must be done in the range of -25°C to -20°C (attempting to calibrate outside this range may cause inaccuracies), the dewpoint must be kept stable during calibration and it must be measured accurately by a reference instrument such as a chilled mirror. One must also note the state of pressure compensation and turn it on or off as applicable to the reference. The necessary steps are described below (also refer to the Set-Up State flow diagram in Appendix B): a. Make sure that the low attenuation and low dewpoint are correctly entered for this sensor. b. SpanCheck™ the sensor. c. Dry-down the sensor for at least 12 hrs. d. Expose the sensor to the known dewpoint (-25°C to -20°C), for a sufficient time (at least 30 min.) e. Go to the CAL/ADJ mode and press the UP button, the display will show the current value, in the range of -1.99 to +1.99 followed by blinking horizontal lines. f. Observe the display, if the buttons are not being pressed; the display will every few seconds show for a short duration the dewpoint as being computed at the moment (utilizing the current cal-adjust value in the computation). g. Use the UP or DOWN buttons to modify the cal-adjust value and observe the alternately displayed dewpoint. Perform this adjustment until the dewpoint matches the value shown by the reference instrument. h. Press the MODE button to go to the next mode, this will save the new cal-adjust value, and it will be retained even if power is turned off.
Entering a known Adjustment Value: If the sensor is being replaced an adjustment value should be supplied with the sensor, to enter it into the instrument:

a. Go to the CAL/ADJ mode and press the UP button, the display will show the current value, in the range of -1.99 to +1.99 followed by blinking horizontal lines, ignore the alternately dis- played dewpoint. b. Use the UP or DOWN buttons to modify the cal-adjust value as necessary. c. Press the MODE button to go to the next mode, this will save the new cal-adjust value, and it will be retained even if power is turned off. 6. Analog Output Span selection 4/20 mA or 0/24 mA: By pushing the UP or DOWN buttons, the user may select the desired span of the analog output. 7. Testing the analog output: By pushing the UP, DOWN, or Pressure Correct buttons, the user forces the analog output to its high, low and mid-range values, respectively. This facilitates the hook-up and testing of the user’s monitoring equipment. 8&9Output range setting: These modes are used to set dewpoints corresponding to the low and/ or high end of the Analog Output. The factory default values are: low end at -100°C, and high end at +20°C. For detailed discussion of the necessary considerations when choosing the ana- log output range, refer to appendix J. 10. Lock/Unlock the instrument with optional SECURED password protection The Lock/Unlock feature of the instrument allows the owner to block access to parameter set- tings, protecting the instrument from unauthorized or inadvertent changes. Attempting to change settings while the instrument is locked will display LOC and beep.

ber can be corrected by reversing the use of the UP and DOWN keys), with the “12” flashing. Pressing the “MODE” button will flash the next “0”. Pressing the “UP” button three times will display “1230”. Pressing the “MODE” button will flash the last “0”. Pressing the “UP” button four times will display “1234”. Pressing and holding the “MODE” button, will display “SET UN/ LOC”, indicating that the instrument is now unlocked. It may be locked without the password, same as any other instrument.
Note: To avoid unauthorized changes, it is imperative that instrument is returned into the locked mode. 11. Sensor Cable Compensation: As noted in other paragraphs this instrument measures the impedance of the sensor (in the form of attenuation in decibels) from which it computes the dewpoint. At very low dewpoints the sensor capacitance is small enough to be in the same order of magnitude as a long coaxial cable. Thus the measurement circuit must be compen- sated for the impedance of the sensor cable. Instruments which are delivered with a cable are already compensated at the factory and should not be compensated again, however if a new cable is connected, the following procedure should be used for compensation (also refer to the Set-Up State flow diagram in Appendix A): a. Disconnect the sensor from the cable by unplugging the BNC connector, keep the cable connected to the instrument. b. Prepare a means by which the BNC connector at the end of the cable could be reliably shorted (using a jumper clip etc.), do not attempt to short at the instrument screw terminals as this approach will not account for the cable resistance. c. Go to the CNF/CBL/OPN (Confirm Cable Open) mode. d. Leave the BNC connector open and press the UP key. The instrument will show OPN and beep for a few seconds. Then it will show CNF/CBL/SHR (Confirm Cable Short). e. Short the BNC connector and press the UP key, the unit will show SHR and beep for a few seconds. Then it will show CNF/CBL/OPN again. f. Leave the BNC connector open again and press the UP button. The instrument will show OPN and beep for a few seconds. Then it will show END
The compensation data will be retained even if power is turned off.

This instrument performs diagnostic tests on power up as well as once every two minutes. The table that follows, depicts all possible error/unexpected messages that may be displayed on the instrument. For each message the table has explanations for the reason, and if necessary a sug- gested action to remedy it. The instrument contains a fuse which may open if the instrument is subjected to voltages exceed- ing the specifications. Under these circumstances the unit display will be blank and the instrument will not draw any current when powered up. The user may replace the fuse with a comparable 0.5A, 250V fast acting fuse, to locate the fuse refer to the XDT mechanical drawing in Appendix E.

LO ERR PROM check sum failed.
HI ERR RAM write/read test failed. cycle power
% ERR Unidentified power-up failure. if problem persists,
C.S.F/ EEPROM Check Sum Failed. return to your representative
ERR/AD A/D converter failure.
for service.
system tests once per 2 min.
ERR/REF Reference voltage for A/D out of spec.
LO/BAT Low supply voltage.
Instrument low temperature range
has been exceeded. make sure that the unit is at a
Instrument high temperature range
temperature of -10°C to +50°C.
has been exceeded.
OPN/ Sensor circuit is open.
SAT/ Sensor is saturated. see table in sensor troubleshooting section 2.5
SHR/ Sensor circuit is shorted.
... Trying to calculate dewpoint for undefined sensor. select sensor and recalibrate.
DB Trying to calibrate an undefined sensor.
see calibration instructions. SEN/2LO Sensor reading is ‘too’ low to be from a saturated sensor,
thus can not be calibrated
2LO While Compensating for cable length, the measurement did not
Check the means of opening & shorting the sensor
cable. see section 3.4.5 - 11 2HI
correspond to the expected open or short values.
ERR/EEP EEPROM write cycle not completed.
if this persists, return for service.
LOC Attempting to modify a locked unit. unlock unit, see set-up state 3.4.5.-10.
SEC Attempting to unlock a secured unit.
HI (flashing while viewing DEWPOINT)
The HI Alarm (alarm #1) relay is deenergized. See section 3.4.4.2 Setting the Alarms.
LO (flashing while viewing DEWPOINT)
The LO Alarm (alarm #2) relay is deenergized.
A.O./OPN The Analog Output ‘sees’ an open circuit check the cables connecting the analog output, make
sure the terminating impedance is less that 500Ω.
PSI flashing when dis- playing dewpoint
Pressure correction is enabled, the dewpoint is computed using
refer to section 3.4.4.1
-70°C (-94°F) with flashing DEWPOINT
An XTR-65 sensor is measuring less than -70°C or -94°F D.P. check instrument upgrade to XTR-100 sensor
RNG number can not be displayed in 3½ digits
XEN Turn on message, Xentaur (Greek Ξ=X)
display is dark No power to instrument check power to instrument, check fuse
Legend: denotes a beeping accompaniment to the message. / denotes alternately flashing messages.

The XDT-PM and XDT-NEMA instrument enclosures may be cleaned when necessary using a damp (moist) cloth.The XDT-OEM bare circuit board may be cleaned by blowing air across the circuit board. The sensor should not be cleaned as it may be damaged.
Spare parts may be ordered through your representative.



0/24 mA range of current for the current loop (0mA=minimum reading, 24mA=maximum reading) 4/20 mA range of current for the current loop (4mA=minimum reading, 20mA=maximum reading) - preferred because less than 4mA can indicate a fault or open circuit, also there is at least 4mA available to power equipment 7 segment type of display configuration - where 7 segments can be used to form the numbers 0 through 9 and most letters A/D analog to digital - a device or circuit that converts an analog voltage
to a digital number absorption retention by penetration into bulk of material adsorption retention as a surface layer on a material accuracy quantification of the magnitude of error of a measurement, or
degree of conformity with a standard. Al2O3 aluminum oxide attenuation the decrease in intensity of a signal, as a result of absorption of
energy, measured in decibels (dB). AWG American Wire Gauge (size of the wire, cross section area of the
wire) baud Bits per second of data transfer (over the serial interface) bit(s) Binary digit(s) - ones and zeros BNC a connector type, used with coaxial cables bypass mounting sensor mounting in a slip stream of sample, which feeds back into
the process stream CE Conformitée Europeane; symbol affixed to a product indicating its
conformance with directives issued by the European Community (EC) CENELEC Comité Europeen de Normalisation Electrotechnique (European
Committee for Electrotechnical Standardization); amongst many others has set standards for intrinsic safety check sum an arithmetic total on a set of data: e.g. the data in the instrument's
memory is summed, and compared to a previously stored sum to check for corrupted data coalescing formation of liquids in gas samples, or separation and collection for
disposal of liquids from gas samples control drawing schematics and system drawing of hazardous area approved appara-
tus with control number issued by approval agency CSA Canadian Standards Association CTS Clear To Send - a label for a RS232 interface control line cUL indicates UL evaluation of a product to Canadian safety require-
ments current loop a circuit where the variation in the current flow is used to indicate
the value of a measurement from the instrument, only 2 wires are
Glossary

used for the circuit (see 4/20) dB see - decibel DB25 a style and size of connector - in this case with 25 pins DB9 a style and size of connector - in this case with 9 pins decibel a dimensionless unit representing the logarithm of the ratio of two
values DEMKO independent test laboratory and the European National Certification
Body for Denmark for the safety testing of electrical products desiccant drying agent, usually made from molecular sieve, silica gel or acti-
vated alumina desorption release of a previously absorbed or adsorbed substance dewpoint saturation temperature of a gas in regards to another gas dewpoint demand mode scheme to control the switching between the desiccant towers of a
dryer based on the actual dewpoint dielectric constant measure of the ability of a material to hold an electric charge when
placed between two metal plates, most commonly measured rela- tive to that of free space Div 2 see Division 2 Hazardous Areas Division 2 Hazardous Areas areas classified as hazardous under abnormal conditions (0.1-10
hrs/year) DSR Data Set Ready - a label for a RS232 interface control line DTR Data Terminal Ready - a label for a RS232 control line EEPROM Electrically Erasable Programmable Read Only Memory - a device
that can remember information, even when it has no power applied to it electromagnetic compatibility compatibility with directives issued under CE or with FCC stan-
dards, concerning electromagnetic interference EMI Electro-Magnetic Interference - external electrical or magnetic sig-
nals from natural or man-made sources that can interfere with a device G/M3 see - grams per cubic meter grams per cubic meter the mass of water vapor present in unit volume of moist air at stan-
dard temperature and pressure. HTF™ Hyper-Thin-Film aluminum oxide sensor technology developed by
Xentaur Corp hygrometer device measuring the amount of water present in a sample hygroscopic property of attracting water or water vapor hysteresis lagging in effect behind it’s cause depending upon whether the con-
dition is approached from above or below the value of interest; pur- posefully programmed into control systems with relays to avoid chatter IGT Research Bulletin 8 Institute of Gas Technology, has developed Charts of the Equilib-
rium Moisture Content of Natural Gases in the Range of -40°F to 250°F and 14.7 to 5000 psi

impedance the ratio of voltage to current at a specified frequency measured in
Ohms, a resistor of the same value would act the same as the mea- sured circuit in-hand micro-climate the warm and moist “climate” created when cupping the sensor in
ones hand. in-situ measurement with sensor located in the process stream intrinsically safe incapable of releasing enough electrical or thermal energy to cause
the ignition of flammable substances listed for a hazardous area classification LBS H2O/mmscf see - pounds of water per million standard cubic feet LCD Liquid Crystal Display - a display that uses a special material (that
is sensitive to electrical voltages) to affect the transmission or reflection of light monograph a systematic learned exposition of a subject, e.g. a chart represent-
ing dewpoint conversions for differing pressures. National Electrical Code a set of US standards related to accepted practices for electrical wir-
ing NFPA National Fire Protection Agency (issuer of the National Electrical
Code (NEC)) NIST National Institute of Standards and Technology NPT National (American Standard) Pipe Taper; an ASTM (American
Society for Testing and Materials) standard of tube fittings with tapered threads parity a way of incorporating error checking into a RS232 data transmis-
sion (on a character by character basis) to check if the data has been received correctly pounds of water per million standard cubic feet . (LBS H2O/mmscf) the mass of water vapor present in unit volume
of moist air at standard temperature and pressure. a measurement unit for moisture, typically used in the natural gas industry ppm parts per million - usually based on volume measure, but must be
stated whether it is by volume or by mass (weight) ppmV parts per million based on volume measure, volume of water vapor
per total volume of gas for an ideal gas. pressure correction computation of dewpoint at a pressure different from the pressure at
the sensor PROM Programmable Read Only Memory - a device that can be pro-
grammed so that data is permanently stored in it (typically used for computer programs) RAM Random Access Memory - a device that holds data as long as it has
power applied to it (the data can be changed with out limit) regenerative dryer purge small flow of dry air from drying column passing through regener-
ating column and desorbing water from desiccant at low pressure response time time it takes for a sensor to react to change in its environment
Glossary

RG58 a type of coaxial cable - the letters/numbers refer to the electrical
characteristic of the cable (resistance/capacitance per foot of length) RG6 A type of coaxial cable - the letters/numbers refer to the electrical
characteristic of the cable (resistance/capacitance per foot of length) RS-232C a specification for electrical signaling (also referred to as RS232)
that describes the voltages, rate of change of the voltages, and the signal names and uses for the signals RTS Request To Send - a label for a RS232 interface control line sample system equipment used to bring conditioned sample to the sensing element;
it may include regulators, filters, heaters/coolers, flow meters, valves and other accessories. saturate reaching maximum holding capacity sintered produced by a process of compacting and heating metal powder,
resulting in the formation of a solid, permeable object. slip stream partial stream extracted from a process stream, of which it is repre-
sentative SpanCheck™ method to check the response of an HTF™ sensor at the upper end
of its measurement range stability a measurement's degree of insensitivity to disturbances. UL Underwriters Laboratories Inc., an organization which evaluates,
tests and certifies products for specific safety situations. uncertainty quantification of the error that can be expected from a measure-
ment, characterizing the range in which the true value can be expected to be in.

XDT User’s Manual Appendices
Power ON initialization and sign on message
30 seconds anywhere (other than "Calibration") without key press
appears when changing settings of locked unit
˚C
DEWPOINT
PSI
Flashing PSI indicates pressure corrected dewpoint
Flashing HI and/or LO
LONG PRESS AND PSI MODE
SET HI
Toggle 14.7 psi
Setting HI Alarm (Relay 1) Setting LO Alarm (Relay 2)
PSI
{
{ {
SHORT PRESS
SET HI
˚C
DEWPOINT
LONG PRESS
{
Displayed only if alarm options are installed
˚C
Toggle 14.7 psi
VALUE
PSI
{
SHORT PRESS
LONG PRESS
SET LO
Cancel Calibration (it is not performed)
Relay polarity: = energize relay above trigger dewpoint = energize relay below trigger dewpoint appropriate polarity selection in conjunction with proper wiring wll provide fail safe operation
˚C
DEWPOINT
Scroll through Alarm activity options
Fault reaction: = faults do Not effect relays = faults "look" like High dewpoint = faults "look" like Low dewpoint = faults Flash the relay
˚C
{
When a third Alarm Relay is installed, setup menu options will appear here similar to Relays 1&2, but with both the HI&LO legends active. This is also the dewpoint alarm for units with resetable Audio-Visual Alarm.
NUMBERS CALCULATING
SET
CALIBRATION
OK
TOO LOW
{ {
see Self-Cal user interface next page
go to Set-Up State
Press together to RESET the instrument
New calibration values are saved
Legend: A slash ' ' in the display area, is used to depict two alternately shown (flashing back and forth) messages.
denotes beeping. Buttons without designator or arrow leading out, perform no function

XDT User’s Manual Appendices
Flow Diagram Operating State User Interface (continued) Manual & Scheduled Self Calibration
Flow diagram items not related to Self Calibration are shaded in gray.
Power ON initialization and sign on message
30 seconds anywhere (other than "Calibration") without key press
appears when changing settings of locked unit
˚C
DEWPOINT
PSI
Flashing HI and/or LO
Flashing PSI indicates pressure corrected dewpoint
{
SET
Confirm Entry into Self Cal Menus
When ALARM OPTIONS are installed menus appear here
Set Reference Gas dewpoint
DEWPOINT
{
values +/-0.1˚C
Set Scheduled Calibration Interval in hours 0 (blank)=Scheduled SelfCal disabled, but SelfCal corrections not affected
View Hours left until next Scheduled Calibration
values +/-1hr.
SET
Begin Self Cal Now
The Self Cal Corrections will not be used in dew computations from now on, until another Self Cal is performed, manually or scheduled. Schedules are not affected.
Cancel Calibration (it is not performed)
disable Self Cal corrections
Confirm performing Self Cal NOW
abort performing Self Cal Now
Scheduled Time Elapsed, Entry Point
During Self Cal the Analog output and Alarm Relays are frozen to their previous states
˚C
dew is stable perform Self Cal
DEWPOINT
DEWPOINT
DEWPOINT
Electric Valve Activation, and 15 min. time out for system stabilization, performed only if electric valve is installed. Displays dewpoint being measured now.
Dewpoint stable testing, 0.5˚C for 3 min. Displays dewpoint being measured now.
Electric Valve Deactivation, and final Dewpoint stable testing, 0.5˚C for 3 min. performed only if electric valve is installed. Displays dewpoint being measured now.
NUMBERS CALCULATING
SET
CALIBRATION
OK
TOO LOW
{ {
go to Set-Up State
Press together to RESET the instrument
New calibration values are saved
Legend: A slash ' ' in the display area, is used to depict two alternately shown (flashing back and forth) messages.
denotes beeping. Buttons without designator or arrow leading out, perform no function

XDT User’s Manual Appendices
Power ON initialization and sign on message
LOC appears when changing settings of locked unit
4 min anywhere without key press
go to Operating State
Choose alternate units to show when displaying dewpoint.
ALT/DSP
blank ˚F ˚C ppm LBS G/m3 PSI % ˚C˚F Temp˚F Temp˚C
Factory default is blank, (there is no alternate display).
XT1 XT6 DB
with the shipped sensor. Do not modify these
settings, unless you are changing the sensor.
XTR-100 XTR-65 measure attenuation of
DB/ -16.24
˚C˚F
Set measured attenuation at a known low dewpoint
changes by 0.01 db
˚C
Set low dewpoint for above attenuation
{
changes by 0.1˚C
SET
if no buttons pressed for 3 seconds shows the dewpoint for a moment.
Calibration Midrange Adjust.
changes by 0.01 units
SET
Displayed only if Analog Output option is installed
Test Analog Output by forcing it to high, low or midpoint.
Hi/LoA.O./OPN
(E. G. if Span is 4 to 20, Hi =20mA, Lo=4mA, Mid=12mA)
A.O./RNG/
Set dewpoint corresponding to low end of analog output
Factory default is low range of sensor, e. g. -100˚C for XTR-100
changes by 0.1˚C
A.O./RNG/
Set dewpoint corresponding to high end of analog output
Factory default is high range of sensor, e. g. +20˚C for XTR-100
changes by 0.1˚C
UN/LOC
Lock or unlock the instrument.
When locked, instrument parameters can not be changed. Factory default is locked.
Cable Length Compensation.
CNF/CBL/OPN CNF/CBL/OPN CNF/CBL/SHR OPN OPN SHR END
Open cable at BNC connector Open cable at BNC connector Short cable at BNC connector
Legend: A slash '/' in the display area, is used to depict two alternately shown (flashing back and forth) messages.
denotes beeping. Buttons without designator or arrow leading out, perform no function

XDT User’s Manual Appendices
~1.45"
flats for 1 " wrench 1 4 /
U
A
76NM
T
LISTED
N
P
H
E

in 100 sintered stainless steel,
14mm x 1.25mm
Removable Sensor encapsulated · Viton A O-ring · 3/4"-16

XDT User’s Manual Appendices


XDT User’s Manual Appendices


XDT User’s Manual Appendices

Standard mounting (with clip-on compression brackets) Environmentally sealed mounting (with gasket and clip-on screw clamp mounting brackets)

XDT User’s Manual Appendices

XDT User’s Manual Appendices
The Sensor is constructed as a capacitor whose dielectric consists of porous Aluminum Oxide as well as the gas that has entered in the pores of the Aluminum Oxide. The plates (electrodes) of this capacitor are an aluminum substrate and a porous gold layer deposited on top of the Alumi- num Oxide, the porous gold electrode allows transfer of gases into or out of the Aluminum Oxide pores. The capacitance due to the Aluminum Oxide is always constant, while the capacitance due to the gas varies according to the gas content and pressure. Since the dielectric constant of water is orders of magnitude larger than that of any gases being measured, the quantity of water vapor present in the pores changes the capacitance of the sensor to a much greater extent than any other system variable. For the same reason (the extremely large dielectric constant of the water mole- cule), any capacitance variations arising from the Aluminum Oxide, such as changes due to tem- perature, are insignificant in relation to the capacitance due to the water content. Thus the sensor capacitance varies greatly in proportion to the water content in the surrounding gas. The sensor is designed such that the relationship of the quantity of water and the resultant capacitance has an "S" shaped curve. At the extremely dry end the curve is asymptotic to the capacitance due to the Aluminum Oxide and the gas, while at the very wet end the curve is asymptotic to the capacitance due to water molecules packed extremely tightly (saturated) in the pores of the Aluminum Oxide. This upper end is an excellent indication of the total pore volume; while the lower end of the curve is an indication of the distance between, & area of the capacitor plates (electrodes), this is the intrinsic capacitance of the sensor. Therefore knowing these two points, a sensor can be cali- brated with compensation for small manufacturing deviations.
As the sensor is used in real world applications, it is exposed to various elements and stress; like any other sen- sor it may over time drift from calibra- tion. However one must note that the intrinsic capacitance will not change with use because it is based simply on the dielectric constant of Al2O3 and the distance between, & area of the plates (electrodes), thus the original factory measured values will always be valid. What may change is the pore volume, due to clogging with contaminants, residual oxidation, metal migration, etc. Thus to re-calibrate the sensor the instrument needs only to “know” the capacitance at the wet end of the curve. As discussed previ- ously the capacitance at the wet end approaches asymptotically a saturated pore capacitance, therefore if the sensor is saturated (the exact water content will not be important since the curve is asymptotic), then the instrument can measure the capacitance and re-calibrate the sensor. This is referred to as SpanCheck™.The advantages of this unique capability of Xentaur HTF™ sensors and instruments are obvious in time and cost savings for re-calibration, as well as ease of sensor replacement.
Saturated Capacitance Asymtote of New Sensor
Saturated Capacitance Asymtote of Aged Sensor
Dry Capacitance Asymtote of New & Aged Sensor
-80 -60 -40 -20 0 20

XDT User’s Manual Appendices
Manufacturers of dewpoint sensors will make unbelievable claims of speed. But without support- ing data revealing the conditions under which the measurements were made, such as flow rate, temperature, the dewpoint at which the test was performed, and most importantly the units of measure. This last aspect can be extremely deceiving, consider the following example. At Xentaur we have performed a variety of tests to determine the response time of ours as well as the competitor’s sensors. For this discussion let’s consider a step change of -40°C to -60°C dew- point, in other words a challenging but typical dry-down. An XTR-100 sensor w/SS sintered filter was placed in an electro-polished stainless steel manifold, the temperature was 23°C and the sam- ple flow rate was 3 LPM. The dewpoint was generated with a computer controlled mass flow valve gas dilution dewpoint generator, and was verified with a NIST and NPL calibrated MBW chilled mirror. The system was allowed to achieve equilibrium at -40°C dp overnight, then the dewpoint was abruptly changed to -60°C, the change is depicted below as time 0. The graph below shows the percent change of recorded dewpoint (°C), the same data is also shown in percent change of the recorded dewpoint converted to ppmV. It is immediately obvious how a sensor that has response times exceeding ten minutes can be claimed to have response time of 3 minutes, simply by manipulating the units of measure. The logarithmic relationship between dewpoint and ppmV makes the ppmV reading appear faster. Since the response time is stated in percent change, the units of measure are automatically hidden from the audience.
% change of ˚C
ement
% change of ppm
erted measur
110 1 2 3 4 5 6 7 8 9 10 20 30 40 50 60 70 80 90 100

XDT User’s Manual Appendices
Of course analyzing on a wet-up step will produce much faster results, because all sensors are fast on wet up, since Al2O3 (like most materials) adsorbs faster than it desorbs. Testing with a small step e.g. in the -10°C to 0°C dewpoint range will produce even faster results, and allow manufac- turers to make far-fetched claims of response time of 5seconds for a 90% of step change. The cus- tomer has to be aware of these manipulations of data and be able to ascertain the real-world response time they can expect out of a sensor. As can be seen from the graph the Xentaur sensor is appreciably faster than any competitors'. This is due to the Hyper Thin Film (HTF) technology, an explanation of the workings of the sensor is provided in theXentaur Hyper-Thin-Film (HTF) Aluminum Oxide Technology notes. We can comfortably make the claim that the Xentaur XTR-100 sensor is faster than any on the market today. To compensate for their sensors' slow response some manufacturers resort to software accelera- tion of response time, even though this scheme provides a seemingly faster response to a single step change, it creates great errors in real systems where multiple changes occur consecutively in differing directions. The sensors' real response lag to consecutive step changes in differing direc- tions is erroneously interpreted as belonging to a single step change, and thus large errors of mea- surement are introduced for long periods of time. An analogy of this situation can be imagined with a defender in a football (soccer) game. A slow player cannot defend by just anticipating the next move of the forward wing; he will simply be faked out by a series of back and forth maneu- vers (consecutive step changes in differing directions). While a fast defender cannot be faked out, he simply follows the forward wing as fast as he moves (Hyper Thin Film).

XDT User’s Manual Appendices
Sample treatment with pre-filtering may be a necessity in many instances. However the filter housing and element may affect the dewpoint reading and/or the response time of the system. Below is a graph comparing measurements from the same calibrated sensor with and without a filter utilizing a polycarbonate bowl. All other parameters, which may affect the measurement, were kept constant. Outside of the sample system, a rather hot and humid environment was cre- ated, to illustrate a worse case condition. As can be seen the filter does not cause problems for dewpoints greater than -50°C. However the systems' ability to make precise measurements quickly deteriorates for dewpoints below -55°C.

at 30˚C temperature and ~100%RH. Test gas flow rate 3LPM. Using XDTsn: 2007 and XTR-100 sensor sn: 18107 · Data with legend "ALL SS" was taken with an all stainless steel system. · Data with legend "Through Filter" was taken with same system but with the addition of a polycarbonate bowl filter
With this in mind the user should select the appropriate filter construction materials. It is impor- tant to note that a stainless steel housing filter will not cause measurement errors, however if the filter element inside the housing is made of hygroscopic fibers, then there will be a delay intro- duced in the measurement.

XDT User’s Manual Appendices
1 AC LIVE
4.3mH
0.1uf
2 AC NEUTRAL
0.0047uf 0.0047uf
5 SIG-IN
6 SIG-RET
7 A-OUT
Sensor casing connects to piping
0.1uf
1M Ohm 33V 400W
8 A-RET
ection
O
structu
Chart recorder, A/D converter etc.
12 PS+
nd lo
Possible Earth Ground at Analog Output monitoring equipment
Logic ground and AC Power ground are connected for safety and electromagnetic interference considerations
connec
16 NO
tion
equipm
grou
LO
RS 232 I/O
PC, Data Logger etc.
Possible Earth Ground at terminating equipment
nd lo
caus
20 NO
connectio
D
n through
equipment gr
ound
1. The sensor ground is isolated from the signal and frame grounds, however it is shunted to frame ground with a 33V Transient Voltage Suppressor,1M Ohm resistor and 0.1uF capacitor to prevent electrostatic buildup and noise pickup. When the sensor is connected to equipment which is grounded to earth ground (because of pipes etc.), there is no ground loop because the sensor is not connected to the earth ground on the circuit board. 2. The optional Analog Output and RS-232 Output are both referenced to the circuit board logic ground, which is jumpered at a single point to the AC power ground. Therefore if these out- puts are connected to equipment which connects the signal grounds to earth ground (this is a common occurrence in PC's), then a ground loop may be formed. This can be best avoided by removing the signal to earth ground connection at the termination equipment; if that is not possible then the AC power ground connection to XDT logic ground may be removed but this may cause a EMI problem. Please consult with your representative for information on optional isolated analog and RS-232 outputs. Keep in mind that in general, a ground loop on a digital line such as RS-232 will not cause problems. 3. The Earth Safety Ground (AC GROUND) at pin 4, should not be omitted since the RFI filter capacitors will cause half the line voltage to develop at the instrument’s ground. This may have safety and EMI implications. 4. When installing DC powered instruments (15 to 30VDC at pins #12 and #11), similar condi- tions may exist, if the DC power source is earth grounded.

XDT User’s Manual Appendices
The current output of the XDT, varies with the measured dewpoint. To use the current to calcu- late the value of the dewpoint measurement, one must know the settings of the low and high ends of the analog output range, then:
D I Z – ( ) H L – ( ) ×
where: I = current supplied by XDT analog output in mA. Z = value of zero output of current loop in mA: for 4/20 = 4, for 0/24 = 0. H = value of High end of Analog Output range, converted to selected engineering units L = value of Low end of Analog Output range, converted to selected engineering units D = dewpoint measured by instrument in selected engineering units. S = span in mA of current loop output: for 4/20 = 16, for 0/24 = 24. Consult section 3.4.5.7&8 and/or appendix B, to check and set the Analog Output low and high ranges; the factory default settings are -100°C and +20°C respectively. For example a unit with factory default settings, supplying 12mA is computed to be measuring a dewpoint of -40°C:
12 4 – ( ) 20 100 – ( ) – ( ) ×
16 100 – ( ) + 40 – =
Note that the computation is such that the current is linear to the selected engineering units. Hence, selecting ppmV or LBS or G/M3 units, will cause the analog output to be linearly propor- tional to those units (approximately logarithmically proportional to dewpoint), refer to the graph that follows. Naturally selecting °C or °F will cause the analog output to be linearly proportional to dewpoint. When monitoring in ppmV or LBS or G/M3, the analog output low & high ranges may have to be adjusted to provide a useful output. Consider an example where the area of interest to be moni- tored is 10 to 100 ppmV, and the analog output is set up with the factory defaults of -100°C to +20°C (which is 0.014 to 23612 ppmV); then the current loop output will vary only from ~4.1 to ~4.2 mA in the area of interest (consult with the graph on the following page). In most instances this would be an unacceptable output for proper monitoring of the measurement. In this example the user should adjust the analog output low & high ranges such that the output range is better suited to the measurement of 10 to 100 ppmV. It may be useful to select the low and high ranges to be 5 and 150 ppmV respectively, thus out of range conditions will be detected properly. Then the low range will be set to 5ppmV which is -65.5°C dewpoint, and the high range will be set to 150ppmV which is -38.5°C dewpoint. Now the current loop output will be 4.55 to 14.48 mA in the range of 10 to 100 ppmV, the ~10mA variation is more than sufficient for a good measurement by the user’s equipment. One may carry out similar calculations for LBS or G/M3 and choose the appropriate settings.While making these computations it may be useful to obtain a copy of Xen- taur’s dewpoint calculator, this is a Microsoft Windows™ program which simplifies the process of converting dewpoint measurement units. It is available at www.xentaur.com If you are not cer- tain how to carry out such calculations, send by e-mail to xentaur@xentaur.com or by fax to (631) 345-5349, your system specifics, and some one will get back to you with appropriate analog out- put settings. In general, if the dewpoint is monitored in °C or °F, there is no need to change the factory default

XDT User’s Manual Appendices
-100°C to +20°C settings, because the 4-20mA provides sufficient resolution to measure the out- put better than the specified accuracy of the sensor.
Relationship of 4-20mA output and instrument reading in ppmV or LBS
Instrument reading in ppmV
10 100 1000 10000
49.3 LBS = -20˚C = 1019.3 ppmV
0.56 LBS = -60˚C = 10.66 ppmV
289.1 LBS = 0˚C = 6063.8 ppmV
High AO range = -38.5˚C
Low AO range = -100˚C High AO range = +20˚C
1.97 LBS = -50˚C = 38.83 ppmV
Low AO range = -65.5˚C
4.5mA
4.1mA
0.1 1 10 100 1000
Instrument reading in LBS of H2O / million cubic feet

XDT User’s Manual Appendices
The instrument uses a simple protocol to communicate to a host computer through its RS-232C interface, see section 3.3.2.5 for electrical connections. The interface is configured as 9600 baud, Even Parity, 8 Bits, 1 Stop; and it echoes all received characters. To illustrate the protocol, the following notation conventions are used in this section: The characters sent or received are depicted in bold. Send refers to sending from the host com- puter to the XDT. Receive refers to receiving from the XDT into the host computer. The com- mands and arguments are shown in capital letters while place holders for choices are labeled with lowercase italic letters, with the allowed characters shown below, “nn” denotes a numeric value. Bytes which are not characters are shown with diagonally placed mnemonics, for example Escape is depicted as ESC, this signifies a single byte (hex 1B). Optional characters or command argu- ments are encompassed with brackets [ ]. The commands may be sent in either upper or lower case. Arguments may be delimited with space or comma.
HELP command Lists all of the available commands. send: H[ELP]C
F
R
Who Reset Login C
F
R
DATA command Selects the reporting format of the current measurement. This formatting will be retained until the next DATA command even if the power is turned off. send: DA[TA] units [interval [mode]]C
R C nn N F G P L G D U
The units argument selects the units in which the measurement will be reported. C,F,P,L & G select the dewpoint in °C,°F,ppm,Lbs H20/mmscf and gm/m3 respectively. D selects decibels which is the measured sensor attenuation. U selects using whatever units are selected by the user on the display.
The interval argument is optional, if it is not entered then the measurement will be reported only on a query with the “?” command. However if it is desired for the unit to report the measurement on a regular interval without a query then an interval in seconds may be specified in the range 1 to 255 seconds. This feature allows the user to log the measurement on a dumb terminal or a printer.

XDT User’s Manual Appendices
The mode argument is optional and may be specified only if an interval was specified. It may be “N” for numeric reporting or “G” for graphic reporting of the measurement. The mode argument defaults to numeric mode. The graphic mode may be chosen only for °C or °F, it will plot a graph of the measurement on printers which can respond to the Epson Graphic Commands, and have at least 80 character columns.
In the NUMERIC mode the instrument will report the measurement in the following format B
P Elapsed Time S
P Alarm Status C
F The bell character is used to denote the beginning of each line. The measurement is a floating point decimal number. The units are the appropriate character string: “degF”, “degC”, “ppmV”, “LbsH2O/mmscf” or “g/m3”. The elapsed time since the unit was powered up (with 24hr roll over) is in HH:MM:SS format. The alarm status appears only if alarms are installed, it is indicated with a character string: HiAlrm or LoAlrm or NoAlrm. If an error condition exists then the report will have the following format: B
B
L ErrorString C
F The ErrorString may be one of the following: “Error SensOpen” or “Error SensShort” or “Error SensSat”.
In the GRAPHIC mode the instrument will transmit Epson Graphic Command compatible data strings at the selected interval. Each string has the following format: DC1
9 DC3 where: DC1 selects the printer; Escape A 01 sets the advance to 1 dot per line; Escape K nnmm selects the graphic mode with mmnn bytes to follow; then the graphic bytes are sent 1 byte per horizontal dot (corresponding to 1 °F or °C being plotted) where the MSbit represents the dot the other bits are not used; Line Feed & Carriage Return advance the line; Escape A 09 sets the advance to the normal 9 dots per line, DC3 deselects the printer. In addition every 200 lines, a text line will be sent to alternately time-stamp the plot or label the extreme dewpoint axis. This will produce a graph which looks as follows.
1 E S C K n
m L
F
R
-100C +20C
17:15:48
m · n

XDT User’s Manual Appendices
The printer should have at least 80 character columns, and it is best to use continuous form paper.
“?” command Requests a single report of the measurement in the currently selected format. send:?C
R The unit will respond in the format selected with the Data command, for example:
F The default factory setting for the instrument is to respond in numeric format in the user chosen units. See the section about the DATA command for more details.
P 01:23:45 S
P NoAlrm C
R
Serial Number Command
F where nn is the instrument serial number
P nn C
R
Who Command
P Ver=n.n C
F where n.n is the instrument software version
Reset Command
R the unit will perform power on reset.
F
R

XDT User’s Manual Appendices
To exchange sensors one must know the measured low attenuation of the sensor at some low dew- point as well as the calibration adjustment value. These numbers can be found on the shipping container serial number label. Through the following steps it is helpful to refer to Appendix A: Flow Diagram of XDT Set-Up State User Interface
1. Power up (or reset) the XDT while holding down the MODE key, in order to enter the Set-Up state. Make sure that the instrument in unlocked, refer to section 3.4.5.-10. 2. Press the MODE key to go to the “Choose Sensor Type” mode. Make sure that the proper sensor type is chosen, e.g. XT1 (for XTR-100) or XT6 (for XTR-65). 3. Press the MODE key to go to the “Set Measured Attenuation @ low dewpoint” mode. The display will show the currently entered attenuation in decibels (dB), this is indicated by hav- ing both the °C and °F on the display. Use the UP or DOWN keys to select the value which appears on the shipping container label. Note that this value is a negative number typically in the range of -18.0DB to -14.0DB. 4. Press the MODE key to go to the “Set low dewpoint” mode. The display will show the cur- rently entered low dewpoint. Use the UP or DOWN keys to select the value which appears on the shipping container label. 5. Press the MODE key to go to the “Calibration Midrange Adjust” mode. The display will alternately show CAL and ADJ. Press the UP key once to see the entered calibration adjustment value. The display will alternately show the currently entered value and the dewpoint being measured, when showing the Cal Adj value the number (in the range of -2.0 to +2.0) will be followed by a blinking “X”, and when showing the dewpoint the display will have the appro- priate legend e.g. °C. Use the UP or DOWN keys to select the value which appears on the shipping container label. 6. Press the MODE key five (5) times to go to the “Lock or Unlock” mode. The display will alternately show UN and LOC. Press the DOWN key to lock the instrument. The display will alternately show only LOC.This prevents inadvertent changes. 7. Press the MODE key once. 8. Now the instrument has stored in it’s EEPROM, the calibration values associated with the new sensor, HOWEVER NOTE THAT IT IS STILL NOT CALIBRATED. The instrument may now be reset. 9. Perform the SpanCheck Procedure in section 3.4.4.3.1. 10. Install the sensor in the sample system and measure. If for some reason the Cal Adj value must be re-entered or modified to improve accuracy the unit does not have to be SpanChecked again.

XDT User’s Manual Appendices
Uncertainty of LBS & ppmV calculations due to +/-3°C measurement accuracy
Calculated LBS H2O / mmscf
Measured Dewpoint °C
Dewpoint -3°C
Dewpoint +3°C
0.7 0.8 0.9 1
7 8 9 10
70 80 90 100
700 800 900 1000
Calculated ppmV

To obtain a RAN fill out this form and fax to (631) 345-5349. this document may be obtained in electronic format at www.xentaur.com and e-mailed back
Customer
Contact Name:
Phone:
Fax:
Process material(s) and/or environments (including radiation) to which the equipment has been exposed. An RAN will not be issued without this information.
OSHA Hazard Communication Standard 29CFR 1910.1200 mandated that we take specific steps to protect our employees from exposure to potential hazards. Therefore, a letter certifying that the equipment has been decontaminated must accompany all equipment exposed to hazardous contamination







XENTAUR CORPORATION, 84F HORSEBLOCK ROAD, YAPHANK, NY 11980 U.S.A. TEL (631) 345-3434 FAX (631) 345-5349
Pages referenced in this manual — manufacturer, support and downloads.
What the fault codes shown on the display mean and how to clear them.
| Code | Meaning | Fix | |
|---|---|---|---|
| 2LO | Check the means of opening & shorting the sensor cable. see section 3.4.5 - 11 | p.37 |
Common problems, likely causes, and what to do.
| Problem | Possible cause | Remedy | |
|---|---|---|---|
| Reading is not changing | Condensation in sample system. | Condensation will occur if the temperature of the sample system, at any point is below (colder) the dewpoint temperature of the sample gas. Once having formed, the sample reaching the sensor will have a dewpoint equal to the temperature of the condensation, regardless of the dewpoint of the source gas. | p.15 |
| Slow Response | 1. Water vapor in the system. 2. Flow rate too low. 3. Sample pipe too large and/or too long. 4. Unsuitable sample pipe. 5. Leaks. 6. Hygroscopic materials in sample system | It is usually more satisfactory to bleed a sample gas at atmospheric pressure through the sensor sampling chamber, and to use 1/8” (3mm) o.d. sample pipe. See below re sample pipe material, also see section 2.3 | p.15 |
| Dry Reading | SpanCheck™, wrongly set, or faulty Sensor. | Verify SpanCheck™, or return sensor for full calibration to your representative. | p.15 |
| Wet Reading | Leak in system or use of unsuitable pipe. | Cure the leak, or replace unsuitable pipe with copper or stainless steel. Flexible con- nections should be made with PTFE pipe. NEVER use rubber or plastic pipe. | p.15 |
| Wet Reading | Comparison of readings with manual cooled- mirror instrument. | This type of indicator reads about 10ºC dry at about -50ºC dewpoint due to tempera- ture gradients within the device. The error increases at drier levels. | p.15 |
| Display Shows SAT | Prolonged exposure to wet gas. | Dry down the sensor, install sensor in either a known dry gas stream i.e. instrument quality air or dry nitrogen, or place sensor in a dry can or bottle of desiccant and seal the container from outside air (the shipping container is designed for this purpose) also see section 3.4.4.3.1 | p.15 |
| Display Shows SHR | 1. Instrument Failure | SHR Disconnect cable from input terminals, if the instrument still reads the problem OPN is with the instrument. However, if the instrument reads then reconnect the cable to the input terminals and check possible causes 2 or 3. | p.15 |
| Display Shows SHR | 2.Short circuit on sensor cable or connections. | SHR Disconnect cable from sensor and if meter still reads , cure the short circuit in the cable or connections or replace cable; otherwise check the sensor. | p.15 |
| Display Shows SHR | 3.Short circuited sensor. | OPN Disconnect cable from sensor and note that the meter reading returns to . Use a new sensor, or apply approximately 20V DC to the sensor MOMENTARILY with the sensor in a known dry condition. Polarity is not important, but the contact MUST be very brief or the sensor may be damaged. Consult with your representative for a cir- cuit diagram. | p.15 |
| Display Shows OPN . | 1. Instrument failure. | Short the SIG and SHIELD contacts of the sensor input terminal, if the instrument SHR reads the problem is in the cable or sensor, otherwise return the instrument for service. | p.15 |
| Display Shows OPN . | 2. Open circuit on cable. | Disconnect cable from sensor and short center pin of plug to the outer shell. If the dis- OPN play still shows , repair cable. | p.15 |
| Display Shows OPN . | 3. Open circuit on sensor. | Check sensor connection or replace sensor. | p.15 |