Rosemount Analytical designs, manufactures and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you MUST properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions MUST be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount Analytical products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.
The information contained in this document is subject to change without notice.
Teflon® is a registered trademark of E.I. duPont de Nemours and Co., Inc. Alconox is a registered trademark of Alconox, Inc. SNOOP® is a registered trademark of NUPRO Co.
WORLD HEADQUARTERS Emerson Process Management Rosemount Analytical Inc. 6565P Davis Industrial Parkway Solon, OH 44139 T 440 914 1261 T 800 433 6076 F 440 914 1262 E gas.csc@emerson.com

PAGE SUMMARY
Cover Updated photo, revision number, and release date
P-6 Removed Compliance section
1-1 Added Mid Range
1-2 Added Mid Range
3-1 Added Mid Range, Backlight over range feature, and a note, using the MID range, set the pressure regulator to 2.5 psig.
3-2 Added a photo of the Signal Board, Added a photo of the Signal Board Test Points, removed the signal board diagnostic jumpers (Figure 3-4)
3-3 Removed the Hi/Lo Range Select Jumper (Figure 3-5), Updated the PPM Range selection switch photo and table, removed the decimal point selection jumpers (Figure 3-7), added Configuration Jumper Settings (Figure 3-7)
3-6 Added step 3 to the Zero Calibration Section (In the Mid Range Configuration, Set the Sample Pressure Regulator to 2.5 psig), Added step 3 to the span cali- bration section (In the Mid Range Configuration, Set the Sample Pressure Regu- lator to 2.5 psig), Added step 6 to the span calibration section (Use R25 and R43 for fine adjustments to the display and recorder output.)
4-3 Updated the Signal Conditioning and Display section to reflect the current PCB’s, updated the Circuit function section.
4-5 Updated the Analyzer Signal Conditioning Block Diagram
6-1 Updated the MATRIX
6-3 Updated the Replacement Parts List
List of Drawings Added Drawing 1-1, 1-2, and1-3
Back Cover Updated Company addresses and contact information
Rosemount Analytical Inc. A Division of Emerson Process Management Contents i


Figure 2-1. Power Supply Board Voltage Select Switches 2-2 Figure 2-2. Temperature Control Board 2-3 Figure 2-3. Cable Gland 2-3 Figure 2-4. Rear View of Model 951C (cover removed) 2-4 Figure 2-5. Rear Panel Wiring Connections 2-4 Figure 3-1. Model 951C Controls, Indicators and Adjustments 3-2 Figure 3-2. Signal Board 3-3 Figure 3-3. Signal Board Test Points 3-3 Figure 3-6. PPM Range Select Switch 3-3 Figure 3-7. Configuration Jumper Settings 3-3 Figure 3-8. Range 2 Selection Jumpers 3-3 Figure 3-9. Power Supply Board 3-5 Figure 4-1. Analyzer Block Diagram 4-4 Figure 4-2. Analyzer Thermal System 4-5 Figure 6-1. Major Assemblies of the Model 951C 6-3 Figure 6-2. Photomultiplier Housing Assembly 6-4 Figure 6-3. Converter Assembly 6-5 Figure 6-4. Case Heater Temperature Control Assembly 6-6
Table 3-1. Resistance of Converter Temperature Sensor vs. Temperature 3-7
Drawing 1-1 Panel Cutout / Installation Drawing Drawing 1-2 Flow Diagram, Lo/Mid Range Drawing 1-3 Flow Diagram, Hi Range
iv Contents Rosemount Analytical Inc. A Division of Emerson Process Management

Rosemount Analytical Inc. A Division of Emerson Process Management Contents v



The purpose of this manual is to provide information concerning the components, functions, installation and maintenance of the 951C NOX Analyzer.
Some sections may describe equipment not used in your configuration. The user should become thoroughly familiar with the operation of this module before operating it. Read this instruction manual completely.
The following definitions apply to DANGERS, WARNINGS, CAUTIONS and NOTES found throughout this publication.
Highlights the presence of a hazard which will cause severe personal injury, death, or substantial property damage if the warning is ignored.
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in injury, death, or long-term health hazards of personnel.
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in damage to or destruction of equipment, or loss of effectiveness.
Highlights an essential operating procedure, condition or statement.
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-1

If this equipment is used in a manner not specified in these instructions, protective systems may be impaired.
To avoid explosion, loss of life, personal injury and damage to this equipment and on-site property, all personnel authorized to install, operate and service the this equipment should be thoroughly familiar with and strictly follow the instructions in this manual. SAVE THESE IN- STRUCTIONS.
Do not operate without doors and covers secure. Servicing requires access to live parts which can cause death or serious injury. Refer servicing to qualified personnel.
This instrument was shipped from factory set up to operate on 115 volt 50/60 Hz. For operation on 230 volt 50/60 Hz, refer to Section 2-3.
For safety and proper performance this instrument must be connected to a properly grounded three-wire source of power.
INTERNAL ULTRAVIOLET LIGHT HAZARD
Ultraviolet light from the ozone generator can cause permanent eye damage. Do not look directly at the ultraviolet source in ozone generator. Use of ultraviolet filtering glasses is recommended.
P-2 Preface Rosemount Analytical Inc. A Division of Emerson Process Management

TOXIC CHEMICAL HAZARD
This instrument generates ozone which is toxic by inhalation and is a strong irritant to throat and lungs. Ozone is also a strong oxidizing agent. Its presence is detected by a characteristic pungent odor.
The instrument exhaust contains both ozone and nitrogen dioxide, both toxic by inhalation, and may contain other constituents of the sample gas which may be toxic. Such gases include various oxides of nitrogen, unburned hydrocarbons, carbon monoxide and other products of combustion reactions. Carbon monoxide is highly toxic and can cause headache, nausea, loss of consciousness, and death.
Avoid inhalation of the ozone produced within the analyzer and avoid inhalation of the sample and exhaust products transported within the analyzer. Avoid inhalation of the combined exhaust products at the exhaust fitting.
Keep all tube fittings tight to avoid leaks. See Section 2-8 for Leak Test Procedure.
Connect rear exhaust outlet to outside vent by a 1/4 inch (6.3 mm) or larger stainless steel or Teflon line. Check vent line and connections for leakage.
Tampering or unauthorized substitution of components may adversely affect safety of this product. Use only factory documented components for repair.
This instrument requires periodic calibration with a known standard gas. See Sections 2-5 and 3-3. See also General Precautions for Handling and Storing High Pressure Gas Cylinders, page P-5.
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-3

The ozone generator lamp contains mercury. Lamp breakage could result in mercury exposure. Mercury is highly toxic if absorbed through skin or ingested, or if vapors are inhaled.
If lamp is broken, avoid skin contact and inhalation in the area of the lamp or the mercury spill.
This instrument’s internal pullout chassis is equipped with a safety stop latch located on the left side of the chassis.
When extracting the chassis, verify that the safety latch is in its proper (counter-clockwise) orientation.
If access to the rear of the chassis is required, the safety stop may be overridden by lifting the latch; however, further extraction must be done very carefully to insure the chassis does not fall out of its enclosure.
If the instrument is located on top of a table or bench near the edge, and the chassis is extracted, it must be supported to prevent toppling.
Failure to observe these precautions could result in personal injury and/or damage to the product.
P-4 Preface Rosemount Analytical Inc. A Division of Emerson Process Management

PRESSURE GAS CYLINDERS
Edited from selected paragraphs of the Compressed Gas Association's "Handbook of Compressed Gases" published in 1981
Compressed Gas Association 1235 Jefferson Davis Highway Arlington, Virginia 22202
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-5
Used by Permission

The following 951C instruction materials are available. Contact Customer Service Center or the local rep- resentative to order.
748214 Instruction Manual (this document)
P-6 Preface Rosemount Analytical Inc. A Division of Emerson Process Management

The following summarized instructions on startup and calibration are intended for opera- tors already familiar with the analyzer.
a. Supply zero gas to rear panel SAM-
PLE inlet and set to 15 psig.
For initial startup, refer to detailed instructions provided in Section 3.
b. Note reading on SAMPLE Pressure
Gauge. It should be the same as in Step 6b. If not, adjust output pres- sure regulator on the zero gas cylin- der as required.
a. Supply upscale standard gas to rear
panel SAMPLE inlet.
b. Note reading on SAMPLE Pressure
Gauge. It should be the same as in Step 7b. If not, adjust output regula- tor on cylinder of upscale standard gas as required.
Supply pressure for sample, upscale standard gas and zero air must be the same. If not, the readout will be in error.
a. Set PPM RANGE Switch (Figure
3-6, page 3-3) for range to be used for sample analysis. Set the front panel RANGE1 Control at normal operating setting, if known, or at about middle or the range if normal setting is not known.
a. Supply sample gas to rear-panel SAM-
PLE inlet at 10 to 17 psig (normally 15 psig).
b. Supply zero gas to rear panel SAM-
PLE inlet.
b. Adjust SAMPLE Back Pressure Regula-
c. Adjust front panel ZERO Control for
reading of zero on meter or recorder.
tor so that SAMPLE Pressure Gauge in- dicates the value appropriate to the desired operating range (normal operat- ing pressure is 3 to 5 psig). Using the Mid range setting, set the sample pres- sure regulator to 2.5 psig See Figure 3-1.
a. Set PPM RANGE Switch (Figure
3-6, page 3-3) at setting appropriate to the particular span gas.
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-7

b. Supply upscale standard gas of accu-
rately known NOX content to rear panel SAMPLE inlet.
It is the responsibility of the user to measure efficiency of the NO2-to-NO converter during initial startup, and thereafter at intervals appropriate to the application, normally once a month.
c. Adjust front panel RANGE1 Control so
that reading on meter or recorder is equal to the know parts-per-million con- centration of NOX in the span gas.
d. Adjust R25 on the signal board so that
the Display value and the recorder out- put are equal.
P-8 Preface Rosemount Analytical Inc. A Division of Emerson Process Management

The case heater assembly of the Model 951C maintains the internal temperature at approximately 50oC (122oF).
The Model 951C NOX Analyzer is designed to measure NOX using one of three sets of ranges designated as Hi, Mid, or Lo. The Hi Range set consists of spans with ranges of 0-100, 0-250, 0-1000, and 0-2500 ppm NOX. The Mid Range set consists of spans with ranges of 0-20, 0-50, 0-200, and 0-500 ppm NOX. The Lo Range set consists of spans with ranges of 0-10, 0-25, 0-100, and 0-250 ppm NOX.
The Model 951C Analyzer has specific ap- plications in the following areas:
The NOX analyzer continuously analyzes a flow- ing gas sample for NOX [nitric oxide (NO) plus nitrogen dioxide (NO2)]. The sum of the concen- trations is continuously reported as NOX.
The analyzer is based on the chemilumines- cence method of NO detection. The sample is continuously passed through a heated bed of vitreous carbon, in which NO2 is reduced to NO. Any NO initially present in the sample passes through the converter unchanged, and any NO2 is converted to an approximately equivalent (95%) amount of NO.
The NO is quantitatively converted to NO2 by gas-phase oxidation with molecular ozone pro- duced within the analyzer from air supplied by an external cylinder. During this reaction, ap- proximately 10% of the NO2 molecules are ele- vated to an electronically excited state, followed by immediate decay to the non-excited state, accompanied by emission of photons. These photons are detected by a photomultiplier tube, which in turn generates a DC current propor- tional to the concentration of NOX in the sample stream. The current is then amplified and used to drive a front panel display and to provide po- tentiometric and isolated current outputs.
To minimize system response time, an internal sample-bypass feature provides high-velocity sample flow through the analyzer.
The Backlight blinks when the analyzer is 10% or more over-range. Selecting a less sensitive (higher) range restores the display function.
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 1-1

LO 0 to 10, 0 to 25, 0 to 100, 0 to 250 ppm NOX MID 0 to 20, 0 to 50, 0 to 200, 0 to 500 ppm NOX HI 0 to 100, 0 to 250, 0 to 1000, 0 to 2500 ppm NOX Repeatability within 0.1 ppm or ±1% of fullscale, whichever is greater Zero/Span Drift less than ±0.1 ppm or ±1% of fullscale, whichever is greater, in 24
hours at constant temperature less than ±0.2 ppm or ±2% of fullscale, whichever is greater, over any 10°C interval from 4 to 40°C (for rate change of 10°C or less per hour) Response Time
(Electronic + Flow) 90% of fullscale in less than 1 minute
Sensitivity less than 0.1 ppm or 1% of fullscale, whichever is greater Detector Operating Pressure atmospheric Total Sample Flow Rate 1 Liter per minute at 20 psig Sample Pressure 138 kPa (20 psig) Ozone Generator Gas U.S.P. breathing-grade air
Ambient Temperature Range 4 to 40°C (40 to 104°F) Analog Output
Potentiometric 0 to +5 VDC, 2000 ohm minimum load Isolated Current Field-selectable 0 to 20 or 4 to 20 mA, 700 ohm max load Display Digital, 4-1/2 digit LCD, readout in engineering units, back-lighted Power Requirements 115/230 VAC ±10%, 50/60 ±3 Hz, 570 W maximum Enclosure General purpose for installation in weather-protected areas Dimensions 8.7 in. x 19.0 x 19.0 in. (H x W x D)
22.0 cm x 48.3 cm x 48.3 cm (H x W x D) Weight 22.2 kg (49 lbs) approximate
Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management 1-2
Ranges

the Power Supply Board (Figure 3-9, page 3-5) position voltage select switches S1, S2, S3 (Figure 2-1, page 2-2) and S3 (located on the Temperature Control Board (Figure 2-2, page 2- 2) in the 230 VAC position.
Carefully examine the shipping carton and con- tents for signs of damage. Immediately notify the shipping carrier if the carton or its contents are damaged. Retain the carton and packing material until the instrument is operational.
Refer to Figure 2-4, page 2-3. On the rear of the analyzer, remove the 6.25 A fuse (P/N 902413) and replace with the 3.15 A fuse (P/N 898587) provided in the shipping kit.
See drawing 654063 for Outline and Mounting dimensions.
Install analyzer in a clean area, free from mois- ture and excessive vibration, at a stable tem- perature within 4 to 40°C.
The power and output (recorder and current) cable glands are supplied loose in the shipping kit to allow cable installation to connectors or terminal strips.
The analyzer should be mounted near the sam- ple source to minimize sample-transport time.
Cable Gland Part No. Power 899330 Recorder 899329
A temperature control system maintains the in- ternal temperature of analyzer at 50°C (122°F) to ensure proper operation over an ambient temperature range of 4°C to 40°C (40°F to 104°F). Temperatures outside these limits ne- cessitate use of special temperature-controlling equipment or environmental protection. Also, the ambient temperature should not change at a rate exceeding 10°C/hr.
Remove rear cover to access terminals. Route each cable through the cable gland and connect to the appropriate connector or terminal strip, tighten the gland.
a. Line Power Connections
Refer to Figure 2-3 (page 2-3), Figure 2-4 (page 2-3), Figure 2-5 (page 2-4) and draw- ing 654063. If this instrument is located on a bench or table top or is installed in a pro- tected rack, panel or cabinet, power may be connected via a 3-wire flexible power cord, minimum 18 AWG (max. O.D. 0.480", min. O./D. 0.270"), through the hole labeled POWER, utilizing connector gland (P/N 899330) provided.
The cylinders of air and span gas should be lo- cated in an area of constant ambient tempera- ture (±10°C).
WARNING
ELECTRICAL SHOCK HAZARD
For safety and proper performance this instrument must be connected to a properly grounded three-wire source of power.
Route the power cable through the cable gland and connect the leads to TB1 on the rear panel. Tighten the cable gland ade- quately to prevent rotation or slippage of the power cable. Since the rear terminals do not slide out with the chassis, no excess power cable slack is necessary.
This instrument was shipped from the factory set up to operate on 115 VAC, 50/60 Hz electric power. For operation on 230 VAC, 50/60 Hz, on
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-3

The following power cord and/or support feet (for bench top use) are available:
If the instrument is permanently mounted in an open panel or rack, use electrical metal tubing or conduit.

Set switch window for voltage required.
Figure 2-1. Power Supply Board Voltage Select Switches

Set switch window for voltage required.
Figure 2-2. Temperature Control Board
2-2 Installation Rosemount Analytical Inc. A Division of Emerson Process Management

b. Potentiometric Recorder Connections
Current recorder interconnection cable specs are as follows:
Refer to Figure 2-3 (below), Figure 2-5 (page 2-4) and drawing 654063. Potenti- ometric recorder connections are made on the rear panel. Route the potentiometric re- corder cable through the cable gland in the hole labeled RECORDER OUTPUT and connect to VOLT OUTPUT terminals.
Potentiometric recorder cable specifications are as follows:
As supplied by the factory, the current output produces a zero of 4 mA. The current output may be adjusted to pro- duce a zero of 0 mA as follows:
c. Current Recorder Connections
Refer to Figure 2-3 (below), Figure 2-5 (page 2-4) and drawing 654063. Current recorder connections are made on the rear panel. Route the current recorder cable through the cable gland in the hole labeled RECORDER OUTPUT and connect to CUR OUTPUT terminals
Figure 2-3. Cable Gland

10 PSI - 17 PSI (70 kPa - 120 kPa)
See Figure 2-5 for connections
Figure 2-4. Rear View of Model 951C (cover removed)
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-3

POWER
OUTPUT
J2 Remote Range Change Connections
RANGE 1
RANGE 2
RANGE 3
RANGE 4
RTN
RTN
+24 VDC IN
+24 VDC IN
CHS GND
CHS GND
TB1
L1/HOT
TB1 AC Power Connections
L2/NEUT
GND
CUR OUTPUT
VOLT OUTPUT
+ - G + -
TB2
TB2 Recorder Connections TB2 Current Output Connections
Figure 2-5. Rear Panel Wiring Connections
2-4 Installation Rosemount Analytical Inc. A Division of Emerson Process Management

2-6 SAMPLE REQUIREMENTS
The instrument requires two gases normally supplied from cylinders. They are:
The sample must be clean and dry before entering the analyzer. In general, before admission to the analyzer, the sample should be filtered to eliminate particles lar- ger than two microns and have a dew point below 90°F (32°C). The factory can provide technical assistance if desired.
a. Air (U.S.P. Breathing Grade)
This is used as both (a) an oxygen source for generation of the ozone required for the chemiluminescence reaction, and (b) a standard gas for zero calibration (nitrogen can also be used). Gas for each purpose must be supplied from a separate cylinder due to different pressure requirements at ozonator and zero inlets.
Proper supply pressure for sample, zero and span gases for the Model 951C is 20 psig (138 kPa).
b. Span Gas
TOXIC AND OXIDIZING GAS HAZARDS
This is a standard gas of accurately known composition, used to set an upscale calibra- tion point. The usual span gas is NO or NO2 in a background of nitrogen.
This instrument generates ozone which is toxic by inhalation and is a strong irritant to throat and lungs. Ozone is also a strong oxidizing agent. Its presence is detected by a characteristic pungent odor.

The instrument exhaust contains both ozone and nitrogen dioxide, both toxic by inhalation, and may contain other constituents of the sample gas which may be toxic. Such gases include various oxides of nitrogen, unburned hydrocarbons, carbon monoxide and other products of combustion reactions. Carbon monoxide is highly toxic and can cause headache, nausea, loss of consciousness, and death.
This instrument requires periodic calibration with a known standard gas. See Section 3-3. See also General Precautions for Handling and Storing High Pressure Gas Cylinders, page P-5.
For maximum calibration accuracy, the concentration of NO in the span gas should be similar to that in the sample gas. Also, the span gas should be supplied to the rear panel SAMPLE inlet at the same pressure as the sample gas. To ensure constant pressure, a pressure regulator may be utilized immediately upstream from the SAMPLE inlet.
Avoid inhalation of the ozone produced within the analyzer and avoid inhalation of the sample and exhaust products transported within the analyzer. Avoid inhalation of the combined exhaust products at the exhaust fitting.
Each gas used should be supplied from a tank or cylinder equipped with a clean, non-corrosive type, two-stage regulator. In addition, a shut-off valve is desirable. Install the gas cylinders in an area of relatively constant ambient temperature.
Keep all tube fittings tight to avoid leaks. See Section 2-8 for Leak Test Procedure.
Connect rear exhaust outlet to outside vent by a 1/4 inch (6.3 mm) or larger stainless steel or Teflon line. Check vent line and connections for leakage.
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-5

The following test is designed for sample pressure up to 5 psig (35 kPa).
For positive assurance that system is leak free, perform one of the tests in Section 5.0.
2-6 Installation Rosemount Analytical Inc. A Division of Emerson Process Management

c. Sample Pressure Gauge
The internal SAMPLE pressure (nominally 4 psig, 28 kPa) is adjusted by rotation of the Sample Pressure Regulator. See Figure 3-1 below.
a. Display
The display is a 4-digit liquid crystal de- vice which always displays NOX concen- tration in parts-per-million. Figure 3-1 below.
Using the MID ranges (20, 50, 200, and 500 ppm), set the sample pressure regulator to 2.5 psig, 17.5 kPa
b. Range Selection
The Model 951C has twelve customer se- lectable ranges, four LO ranges (10 ppm, 25 ppm, 100 ppm and 250 ppm), four MID ranges (20 ppm, 50 ppm, 200 ppm and 500 ppm), and four HI ranges (100 ppm, 250 ppm 1000 ppm and 2500 ppm). The range is selected by positioning the PPM RANGE Switch (SW1) and the five con- figuration jumpers on the Signal Board to the desired range controlling the recorder output. Refer to Figure 3-2, page 3-2.
d. Ozone Pressure
The OZONE pressure is determined by the pressure regulator of the air supply cylinder. A nominal pressure of 20 to 25 psig (138 to 172 kPa) is recommended. Proper operation is indicated when the front panel OZONE indicator lamp is lit.
If ozone lamp does not light, increase pressure slightly by adjusting pressure regulator control on the air cylinder.
The Backlight will blink for values 10% in excess of the range maximum. Moving the switch to the right selects a higher fullscale value and restores the display.
POWER SUPPLY BOARD (See Figure 3-3)

CASE HEATER TEMPERATURE CONTROL ASSEMBLY (See Figure 6-4) TEMPERATURE CONTROL BOARD (See Figure 2-2)
Ozone Indicator Lamp (Signal Board DS2) Range1 Control (Signal Board R101)
Figure 3-1. Model 951C Controls, Indicators and Adjustments
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-1

e. Zero and Span Potentiometers
f. Ozone Interlock
See Figure 3-1 (page 3-1) and Figure 3-2 below. Screwdriver access holes through the front panel allow adjustments of the ZERO, RANGE1 and RANGE2 potenti- ometers (R100, R101 and R102 on the Signal Board).
The ozone-producing UV lamp will not ig- nite or stay lit unless adequate air pres- sure is present at the AIR inlet (see Figure 2-4, page 2-3). Nominal setpoint pressure is 20 to 25 psig.
PPM Range Select Switch (Figure 3-6, page 3-3) Lo/Mid/Hi Selection Jumpers (Figure 3-6, page 3-3)

Figure 3-2. Signal Board

Figure 3-3. Signal Board Test Points
3-2 Operation Rosemount Analytical Inc. A Division of Emerson Process Management

c. Set Range2 Selection jumpers to
The following are detailed instructions on startup and calibration.
Range 4, 2500ppm as shown in Figure 3-8 below.

JP1 10 20 100 JP2 25 50 250 JP3 100 200 1000 JP4 250 500 2500
a. Using a small flat screwdriver, set
PPM RANGE Switch (Figure 3-6 below) to the range that will be used for sample analysis.
Figure 3-8. Range 2 Selection Jumpers

Hi Range 1 10 20 100 2 25 50 250 3 100 200 1000 4 250 500 2500 Other Remote Remote Remote
Figure 3-6. PPM Range Select Switch
b. Set Configuration Jumper Settings to
desired position as shown in Figure 3-7 below.

Figure 3-7. Configuration Jumper Settings
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-3

b. Adjust SAMPLE Backpressure Regula-
tor so internal SAMPLE Pressure Gauge indicates the value appropriate to the desired operating range.
a. Adjust OZONE Pressure Regulator so
that OZONE Pressure Gauge indi- cates 20 to 25 psig (138 to 172 kPa).
Inability to obtain a flow of one liter per minute at the EXHAUST outlet usually indicates insufficient sample supply pressure at the SAMPLE inlet. Use a 2400 cc flowmeter (i.e., Brooks P/N 1350) at the EXHAUST outlet to measure flow.
b. To establish correct pressure of zero gas,
supply zero gas to rear panel SAMPLE inlet. Note reading on internal SAMPLE Pressure Gauge. It should be the same as the nominal 4 psig (28 kPa) for the Lo and Hi Ranges and 2.5 psig for the Mid Range SAMPLE pressure indicated on the internal SAMPLE pressure gauge. This should remain constant when the analyzer input SAMPLE is switched from calibration gas standard to a zero gas standard. This may be assured by setting the delivery from the SAMPLE and the zero gas cylinder of span gas cylinder to the same value of delivery pressure, nominally 20 psig (138 kPa). If not, ad- just output pressure regulator on zero gas cylinder as required.
a. Supply upscale standard gas to rear
panel SAMPLE inlet.
b. Note reading on internal SAMPLE
Pressure Gauge. It should be the same as in Step 3b.
Supply pressures for sample and upscale standard gases must be the same. Otherwise, readout will be in error.
a. Supply sample gas to rear panel
SAMPLE inlet.
The analyzer is now ready for calibration.
3-4 Operation Rosemount Analytical Inc. A Division of Emerson Process Management





Figure 3-9. Power Supply Board
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-5

a. Zero Calibration
After calibrating analyzer per Section 3-3, supply sample to SAMPLE inlet. Set PPM RANGE Switch (Figure 3-6, page 3-3) in ap- propriate position. The instrument will now continuously analyze the sample stream.
b. Upscale Calibration
The Model 951C is designed for continuous operation. Normally, it is never turned off ex- cept for servicing or for a prolonged shut- down.
During periods of shutdown, turn off the ozone lamp by shutting off the input air source.
Once the appropriate high voltage and elec- tronic gain have been selected such that the named calibration gas value is indicated by the Model 951C, the instrument is ready for adjustment of the converter temperature.
The vitreous carbon converter used in this analyzer has a low surface area which gradu- ally increases during high temperature opera- tion of the converter material.
3-6 Operation Rosemount Analytical Inc. A Division of Emerson Process Management

Initially, the temperature of the peak of the converter efficiency starts at a relatively high value because significant heat must be sup- plied to make the converter active enough to reduce the input nitrogen dioxide to nitric ox- ide at the required 95% level. During the op- eration of the analyzer, the temperature of the peak will fall as the surface area of the con- verter is increased and less external energy is required to cause adequate conversion.
4. Increase the temperature of the converter by rotating the converter temperature ad- just potentiometer (R9 CONV HTR) one quarter turn clockwise, wait fifteen min- utes for thermal equilibrium and then re- measure the NO2 calibration gas value. Note its value. Repeat this procedure of one quarter turn adjustments of the poten- tiometer, waiting for thermal stability and determination of the calibration gas value until either a 95% value is obtained or the final one quarter turn adjustment gives an efficiency increase of less than one per- cent.
In extreme cases, where converter re-profiling has not been conducted, the converter is so active that it not only reduces nitrogen dioxide to nitric oxide, but it reduces the nitric oxide produced to nitrogen, which is not detected by the chemiluminescence reaction. The remedy in this case is to adjust the converter tempera- ture to a lower value to improve the converter efficiency.
It is important that the converter temperature be periodically profiled to assure that it is run- ning at its peak efficiency. An interval of one week is recommended. The nominal range of operational temperatures for the converter is 275°C to 400°C (527°F to 750°F). The operat- ing temperature of the converter may be con- veniently checked on the Power Supply Board (Figure 3-9, page 3-5) by momentarily de- pressing switch CONV TEMP CHECK (S4) while monitoring the resistance across termi- nals TP1 and TP2.
Table 3-1 allows for conversion of the ob- served resistance to the operating tempera- ture for the converter.
RESISTANCE
( ° C)
(Ohms) 0 400 25 438 100 552 200 704 250 780 300 856 350 932 400 1008 450 1084
Follow this procedure to optimize the operat- ing temperature of the converter:
Table 3-1. Resistance of Converter Temperature
Sensor vs. Temperature
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-7

Converter temperature is not a direct measure of converter efficiency. Temperature measurement is for reference purposes only.
3-8 Operation Rosemount Analytical Inc. A Division of Emerson Process Management

consume nitrogen dioxide as it delivers sample to the analyzer. Losses of nitro- gen dioxide in the pump will be reported as converter inefficiency.
It is the responsibility of the user to measure efficiency of the NO2-to-NO converter during initial startup, and thereafter at intervals ap- propriate to the application (normally once a month).
Section 3-5 optimizes the operating tempera- ture of the converter. It also serves as an effi- ciency check if the concentration of NO2 in the calibration gas is documented accurate rela- tive to National Institute of Standards and Technology (NIST) Reference Materials. If the concentration of the nitrogen dioxide calibra- tion gas is not known accurately, this proce- dure still serves to adequately provide the correct converter operating temperature.
5. If the response at the end of the thirty mi- nute period decreases more than 2.0 per- cent of the highest peak value observed, the system is not acceptable and correc- tions must be made before repeating the check. If it is determined that observed subnormal conversion efficiencies are real, and not due to errors introduced by nitrogen dioxide consumption in the sam- ple pump or other parts of the sample handling system, verify that the converter is peaked at the optimum temperature be- fore replacing with a new converter.
If the only available known standard is the ni- tric oxide calibration standard, the following procedure may be performed. This procedure checks converter efficiency through the utiliza- tion of gas-phase oxidation of nitric oxide into nitrogen dioxide over a range of nitrogen diox- ide concentrations. This technique is ab- stracted and adapted from 40 CFR, Pt. 60, App. A, Method 20, Paragraph 5.6.
After initial startup or startup following a shut- down, the analyzer requires about two hours for stabilization before it is ready for calibra- tion. Maximum permissible interval between calibrations depends on the analytical accu- racy required, and therefore cannot be speci- fied. It is recommended that initially the instrument be calibrated at least once every 8 hours. This practice should continue until ex- perience indicates that some other interval is more appropriate.
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-9



THEORY
The flow rate of the selected gas into the reaction chamber is controlled by a back pressure regulator inside the ana- lyzer. It provides an adjustable, con- trolled pressure on the upstream side, where gas is supplied to the calibrated, flow-limiting sample capillary. The regu- lator is adjusted for appropriate reading on the internal SAMPLE Pressure Gauge. For operation at NO and NO2 levels below 250 ppm, correct setting on the SAMPLE Pressure Gauge is 4 psig (28 kPa). This results in a flow of approximately 60 to 80 cc/min to the re- action chamber.
The chemiluminescence method for detection of nitric oxide (NO) is based on its reaction with ozone (O3) to produce nitrogen dioxide (NO2) and oxygen (O2). Some of the NO2 molecules thus produced are initially in an electronically excited state (NO2*). These revert immediately to the ground state, with emission of photons (essentially red light).
The reactions involved are:
NO + O3 → NO2* + O2
NO2* → NO2 + Red Light
Excess sample is discharged with the effluent from the reaction chamber via the EXHAUST outlet. Bypass flow is set by the restrictor at 1 L/min (nominal) to ensure proper functioning of the SAM- PLE Pressure Regulator and rapid sys- tem response. Excessive changes, on the order of 5 psig (35 kPa), in the pressure of the sample or standard gas will affect the bypass flow rate and can affect accuracy.
As NO and O3 mix in the reaction chamber, the intensity of the emitted red light is proportional to the concentration of NO.
(Any NO2 initially present in the sample is re- duced to NO by a heated bed of vitreous carbon through which the sample is passed before be- ing routed to the reaction chamber.)
The intensity of the emitted red light is meas- ured by a photomultiplier tube (PMT), which produces a current of approximately 3 X 10-9 amperes per part-per-million of NO in the reac- tion chamber.
b. Ozone Generation
Suitably pressurized air from an exter- nal cylinder is supplied to the rear panel AIR inlet. The proper pressure setting is 20 to 25 psig (138 to 172 kPa). Within the ozone generator, a portion of the oxygen in the air is converted to ozone by exposure to an ultraviolet lamp. The reaction is:
The analyzer flow system is shown in drawing 2-0 and 3-0. Its basic function is to deliver regu- lated flows of sample, calibration gas, or zero gas and ozonized air to the reaction chamber. The discharge from the reaction chamber flows from the analyzer via the EXHAUST outlet.
From the generator, the ozonized air flows into the reaction chamber for use in the chemiluminescence reaction.
a. Flow of Sample, Standard Gas or Zero
Gas to Reaction Chamber
Suitably pressurized sample, standard gas or zero gas is supplied to the rear panel SAMPLE inlet.
Rosemount Analytical Inc. A Division of Emerson Process Management Theory 4-1

cluded jumpers, this is used for higher range selection (ie) 100 to 2500 ppm NOx.
A block diagram of the signal-processing elec- tronics is shown in Figure 4-1, page 4-3.
Gain Amplifier U14 The precision amplifier U14 allows 2 se- lectable gains, on for 10, 100, and 1000 spans and the other for 25, 250, and 2500 spans.
The signal conditioning and display board pro- vides the following functions:
Intermediate gain ranges as required by the 951C Lo Range: 10, 25, 100, 250 ppm NOx, Mid Range: 20, 50, 200, 500 ppm NOx or Hi Range 100, 250, 1000, 2500 ppm NOx are obtained by interpo- lating between the requisite ranges us- ing Boolean Logic and Analog Switches.
Range 1 and Range 2 selection The signal output of U14 is adjusted at the front panel for instrument calibration purposes. Any second range may be chosen by placing a jumper on JP1, JP2, JP3, or JP4. When the pro- grammed range is selected, the K2 re- lay toggles and allowing that range to be calibrated. Note! If no jumper is pre- sent only range 1 may be calibrated
All of the above functions are on a single board located at the front of the instrument. Certain control requirements (ie) range calibration and zero offset adjustment are available as screw- driver adjustments via the front panel. A digital liquid crystal display is mounted on the rear side of the board for data display purposes as are control potentiometers.
Post Amplifiers U5 The conditioned signal from U14 and the range calibration potentiometers R101 and R102 are provided to U5. This amplifier has switch controlled feed back resistors that permit gain selec- tions of any range, as determined by the range switch and associated logic. An Adjustment potentiometer R25 al- lows a small correction for inter range calibration purposes.
a. Circuit Functions
Signal conditioning amplifiers Boards assembly 6A00326G01/02 is a high input impedance electrometer amplifier board.
The output of U14 is provided to the Analog to Digital converter for digitiza- tion and display purposes. Potentiome- ter R43 allows adjustment of the signal to the ADC allowing concordance of dis- play and recorder signals if required. Normally the ADC should be correctly calibrated during the factory test proce- dure and should not require further ad- justment. Potentiometer R25 allows
Current output from the detector unit is con- verted to voltage by the electrometer ampli- fier and then further amplified by post amplifier.
The electrometer amplifier gain may be re- duced by a factor of 10 by shorting JP1 and by shorting pins 1 and 2 on P1 with the in-
4-2 Theory Rosemount Analytical Inc. A Division of Emerson Process Management

recorder output adjustment to match dis- play.
Connecting any input terminal 1 thru 4 to ground terminals 9 or10 will select one of the four ranges.
U5 is the recorder output amplifier. The out- put signal may be trimmed by R25 to pre- cisely match the recorder calibration. Note that this adjustment does not affect or change the instrument display,
For remote operation a 24V power sup- ply should be connected to terminals 5,6 and 7,8, negative or low side is connected to 5, 6. Positive or high side to 7,8. For protection purposes the high side is fused. Note! The local range switch SW1 must be in the fifth posi- tion (position Five) to disconnect lo- cal control and provide the return for remote control.
Analog to Digital Converter U15 is a special integrated circuit with a built in ADC and LCD Drivers. The signal data is digitized and provided in the correct format to the liquid crystal display (LCD).
For local operation Set the range switch SW1 between setting 1 thru 4. An ex- ternal 24V supply is not required. SW1 is located on the top of the signal board.
Display/Backlight and over range blink Circuits The display circuits comprise of U15, no ad- justments are required for these functions
The blink or over range display function uses U5, U6, and U9.
The Analyzer Thermal System is shown in Figure 4-2, page 4-3. Its basic function is to provide a stable thermal environment for the PMT.
U5 is an analog comparator that senses an over range condition and turns on U6. U6 sets the blink rate for the Backlight.
The temperature of the PMT must be held within a half-degree band at approximately 18°C if it is to produce a useful signal for low concentrations of NOX. This is accom- plished by means of a solid-state cooler which houses the PMT. The heat which is radiated from the cooler is carried away by the Cooler Fan.
Remote Control Circuits The Model 951C is provided with a fully iso- lated remote control interface. Optical isola- tors and a remote 24V power supply ensure that no direct return path exists between the users system and the 951C When in the remote mode of operation.
Relay K1 switches between the internal range switch control (SW1) and the External Remote Range Control. When SW1 is on setting 5 it causes K1 to toggle and a fully isolated system is achieved.
The solid-state cooler must work against a relatively constant load in order to maintain the temperature of the PMT. This load is produced by a case heater and exhaust fan which control the temperature inside the case within a one-degree band (approxi- mately 50°C for ambient temperatures from 4°C to 40°C).
Optical Isolators U7, U8, U10, and U13 con- nects via a ribbon cable to the remote con- nector panel at the rear of the 951C. A ten terminal barrier strip provides connection for remote range selection.
The electronics which support the Analyzer Thermal System and the NO2-to-NO Con- verter are contained on the Power Supply Board.
Rosemount Analytical Inc. A Division of Emerson Process Management Theory 4-3

PMT Electrometer Gain Amplifier Span Amplifier
Span Adjustments R101, R102
Figure 4-1. Analyzer Block Diagram
Rosemount Analytical Inc. A Division of Emerson Process Management Theory 4-5

Top View of Analyzer
SOLID-STATE COOLER
PMT
CASE HEATER
FRONT PANEL
Figure 4-2. Analyzer Thermal System
4-6 Theory Rosemount Analytical Inc. A Division of Emerson Process Management
Cooling Fins · Cooler Fan · Fan Heater

Rosemount Analytical Inc. A Division of Emerson Process Management Theory 4-7

a. Display Fullscale Span Adjustment
If a recorder is used, and has been prop- erly zeroed, it should agree with the dis- play reading. If not, obtain agreement by adjustment of R25 on the Signal Board (see Figure 3-1 (page 3-1) and Figure 3-2 (page 3-2). If agreement cannot be reached, check the recorder. If the re- corder is functioning properly, replace the Signal Conditioning Board.
Servicing requires access to live parts which can cause death or serious injury. Refer servicing to qualified personnel.
INTERNAL ULTRAVIOLET LIGHT HAZARD
Ultraviolet light from the ozone generator can cause permanent eye damage. Do not look directly at the ultraviolet source in ozone generator. Use of ultraviolet filtering glasses is recommended.
b. Overall Sensitivity
Principal factors that determine overall sensitivity of the analyzer are the follow- ing: (a) sample flow rate to the reaction chamber, (b) sensitivity of the photomulti- plier tube (PMT), and (c) PMT high volt- age. If specified fullscale readings are unobtainable by adjustment of the SPAN Control, sensitivity is subnormal. The cause of reduced sensitivity may be in ei- ther the flow system (See Section 5-2, page 5-3) or the electronic circuitry (See Section 5-6, page 5-10).
The photomultiplier tube must not be exposed to ambient light. If the photomultiplier tube is ex- posed to light while the power is on, either through a loose fitting on the reaction chamber or any other leak, it will be destroyed. If exposed to ambient light with the power off, the tube will be noisy for some period of time. Unless appro- priate precautions are observed, light can strike the tube upon removal of fittings from the reac- tion chamber. 5-1 SYSTEM CHECKS AND ADJUSTMENTS
If either the High Voltage Board or the Phototube/Reaction Chamber Assembly has been replaced, a readjustment of the high voltage will probably be required to obtain the correct overall sensitivity. Ad- just R30 on the Power Supply Board (see Figure 3-1 page 3-1 and Figure 3-9 page 3-5) clockwise to increase (negative) the photomultiplier high voltage and sensitiv- ity, or counterclockwise to decrease (negative) the voltage and sensitivity. The adjustment range is about -650 V to -2100 V for the regulated DC voltage ap- plied to the photomultiplier tube. Nominal setting is -1100 volts. However, the volt- age should be adjusted as required for overall system sensitivity.
The following procedures may be used to de- termine the cause of unsatisfactory instrument performance, or to make adjustments follow- ing replacement of components. If a recorder is available, use it for convenience and maxi- mum accuracy in the various tests.
Rosemount Analytical Inc. A Division of Emerson Process Management Troubleshooting 5-1

c. Ozone Output
display reading with adjustment of the ZERO Control. If this cannot be accom- plished, the cause must be found and cor- rected. The fault may be in either the electronic circuitry or the sample flow sys- tem.
TOXIC GAS HAZARD
Use extreme caution in troubleshooting the ozone generator. Ozone is toxic.
First, establish proper performance of the electronic circuitry. Turn on analyzer power. Verify that ZERO Control and am- plifier are functioning properly. Then, check for excessive photomultiplier dark current and/or contamination of the reac- tion chamber or sample flow system as follows:
To check for adequate output from the ozone lamp, a convenient technique is to calibrate the analyzer on a high level NO standard such as 250 ppm NO at the nominal 4.0 psi internal sample pressure setpoint, and note the reading. The sam- ple pressure setpoint is then sequentially set to pressures of 3.0, 2.0, and 1.0 psi after a stable span gas reading has ob- tained at the higher pressure setpoint. The span gas value will change as the pressure is changed. The difference in span gas value between two successive sample pressure levels should be ap- proximately the same for the 4.0 to 3.0, 3.0 to 2.0, and 2.0 to 1.0 pressure steps.
To check, shut off all flow to the ozone generator. Turn off ozone generator. Supply cylinder air to rear panel SAMPLE inlet. Note response on display or recorder. If back- ground is still excessive, possible causes are:
If the size of the span gas value differ- ence increases as the sample pressure is lowered, the analyzer output is limited by the amount of ozone production from the lamp and the two additional checks should be made. First, verify that the sample flow (not including bypass) does not exceed the nominal 60 to 80 cc/min, at 4.0 psi internal sample pressure. Sec- ond, substitute another lamp to see if the ozone output is increased.
leakage of ambient light to photomul- tiplier tube
electrical leakage in socket assembly
See Section 5-4a, page 5-7.
If no other ozone lamp is available, the analyzer sample input pressure may be reduced to the pressure where the ozone limitation is not present. If the lamp output is low and the sample pressure is reduced to restore operation to the condition where ozone limitation is not occurring, some degradation in analyzer response time characteristics may occur.
To facilitate servicing and testing, the Model 951C has front drawer access.
Drawing 654090 shows flow system details, including fittings, thread specifications and connecting tubing.
d. Background Current
a. Cleaning Sample Capillary
With zero air supplied to rear panel SAM- PLE inlet, excessive background current is evidenced by the inability to obtain zero
If clogging of sample capillary is sus- pected, measure flow rate as described below.
5-2 Routine Servicing Rosemount Analytical Inc. A Division of Emerson Process Management
defective photomultiplier tube

Do not over-tighten capillary internal fitting, as over-tightened fittings may restrict the sample flow.
b. Ozone Restrictor Fitting
With instrument power off, supply suitable test gas (dry nitrogen or air) to rear panel AIR inlet. Cover photomultiplier housing with a dark cloth. At the fittings on the re- action chamber, disconnect the ozone tube and place a cap over the open fitting to prevent entry of ambient light. Connect a flowmeter to open end of ozone tube. Adjust the OZONE Pressure Regulator so that the OZONE Pressure Gauge indi- cates normal operating pressure of 20 to 25 psig (138 to 172 kPa). Verify that test flowmeter indicates an appropriate flow of 500 to 600 cc/min for 20 psig.
If the opened fitting is inadvertently exposed to ambient light, the instrument will temporarily give a highly noisy background reading. If so, this condition may be corrected by leaving the instrument on, with high voltage on, for several hours. If high voltage is on during exposure, the photomultiplier tube will be destroyed.
Subnormal flow indicates clogging in the flow path that supplies air to the ozone generator. This path contains a Restrictor (P/N 655519), consisting of a metal fitting with internal fritted (metal membrane) re- strictor to reduce pressure. The fitting is upstream from the inlet port of the ozone generator. If the internal restrictor be- comes plugged, the assembly (P/N 655519) must be replaced as it cannot normally be cleaned satisfactorily.
This assembly consists of the photomultiplier tube and socket, the thermoelectric cooler, and the reaction chamber. Refer to Figure 6-1 (page 6-4) for location and details of mount- ing. Refer to Figure 6-2 (page 6-5) for infor- mation on the assembly.
The assembly must be removed from the ana- lyzer in order to clean the reaction chamber or to replace the photomultiplier tube.
Rosemount Analytical Inc. A Division of Emerson Process Management Routine Servicing 5-3

a. Removal
If sample gas is properly filtered, the reac- tion chamber should not require frequent cleaning. In event of carryover or con- tamination, however, the chamber should be disassembled to permit cleaning the quartz window and the optical filter. The following procedure is recommended.
To remove the photomultiplier tube/reaction chamber assembly from the analyzer, do the follow:
Always wear surgical rubber gloves when handling the reaction chamber to prevent contamination from handling.
b. Cleaning Reaction Chamber
Photomultiplier tube will be permanently damaged if exposed to ambient light while powered with high voltage. Photomultiplier tube will develop temporary electronic noise if exposed to ambient light with high voltage OFF. A temporary noisy condition may be corrected by leaving instrument on, with high voltage on, for several hours. The required recovery time depends on intensity and duration of the previous exposure. Noise level on the most sensitive range usually drops to normal within 24 hours.
Using a stiff plastic bristle brush, such as a toothbrush, scrub the Teflon surface and gas ports of the reaction chamber with clean distilled water and Alconox* de- tergent (P/N 634929). Alconox detergent is included in the shipping kit provided with the Model 951C NOX Analyzer, and is available from Sargent-Welch Scientific
5-4 Routine Servicing Rosemount Analytical Inc. A Division of Emerson Process Management
Standard Cleaning Procedure

Company under its catalog number S-195650-A.
c. Photomultiplier Tube and Housing
The photomultiplier tube operates at high DC voltages (nominal setting is -1100 volts) and generates small currents that are highly amplified by the sig- nal-conditioning circuitry. It is therefore important that ambient humidity and con- densed water vapor be excluded from the interior of the photomultiplier housing. Ambient humidity can result in electrical leakage, observed as abnormally high dark current. Water vapor or condensed moisture in contact with the photomulti- plier tube may result in an abnormally high noise level during instrument readout on zero air or upscale standard gas.
Using Alconox and clean, soft facial tissue (NOT industrial wipes), carefully clean the quartz window. Vigorously flush reaction chamber and quartz window with clean distilled water. Blow out all possible water from internal passages of reaction cham- ber. Dry reaction chamber and quartz window in a warm oven at 125°F to 150°F (52°C to 66°C) for 30 to 45 minutes or purge-dry the parts with dry cylinder air or nitrogen to eliminate all moisture.
ACID HAZARD
The Photomultiplier Tube/Reaction Chamber Assembly incorporates several features for exclusion of humidity and moisture. The photomultiplier socket as- sembly is potted with high impedance sili- cone rubber compound and is sealed from external influences with epoxy and rubber gasket material. The socket as- sembly and the reaction chamber are sealed with O-rings into opposite ends of the tubular photomultiplier housing. The socket end of the housing may be sealed with either one or two O-rings, depending on the length of the phototube.
Hydrochloric acid (HC1) is a strong acid. It is irritating to the skin, mucous membranes, eyes and respiratory tract. Direct contact causes severe chemical burns.
Avoid Contact with eyes and skin and avoid breathing fumes. Use in hood or well ventilated place. Wear goggles, rubber gloves and protective clothing.
Alternate Cleaning Procedure - For High Residual Fluorescence
Holding the reaction chamber by the tube fittings, and using appropriate caution, immerse the white Teflon part of the chamber in 50% concentrated Reagent Grade hydrochloric acid. After five min- utes, rinse thoroughly with de-ionized wa- ter, then air dry as in the standard cleaning method above.
d. Replacement of Photomultiplier Tube
The photomultiplier tube assembly must be removed from the housing in order to replace the tube. To remove, do the fol- lowing:
Place parts in position and press on end-cap so that mating threads engage properly, without cross threading. Turn mating parts in one continuous motion un- til the parts mesh. Do not over-torque.
With reaction chamber now assembled, replace and reconnect it in reversed re- moval sequence. Orient as noted in step 2.
Rosemount Analytical Inc. A Division of Emerson Process Management Routine Servicing 5-5

a. Lamp/Housing Removal
To remove the lamp and housing, do the follow:
This system consists of the ultraviolet lamp, lamp housing, and power supply. Refer to Figure 6-1 for location and details of mount- ing.
The ozone generator lamp contains mercury. Lamp breakage could result in mercury exposure. Mercury is highly toxic if absorbed through skin or ingested, or if vapors are inhaled. Handle lamp assembly with extreme care.
If lamp is broken, avoid skin contact and inhalation in the area of the lamp or the mercury spill. Immediately clean up and dispose of the mercury spill and lamp residue as follows:
b. UV Lamp Replacement
To replace the lamp, do the following:
Do not touch lamp. Fingerprints may cause a decrease in lamp output.
Do not place in trash, incinerate or flush down sewer. Cover any fine droplets of mercury in non-accessible crevices with calcium polysulfide and sulfur dust.
Replace the lamp and housing by revers- ing the steps in this section.
5-6 Routine Servicing Rosemount Analytical Inc. A Division of Emerson Process Management

c. Power Supply Removal
than the outlet tube. Further, the outlet tube and the sample capillary (P/N 615989) connect to the same stainless steel tee.
To remove the Power Supply, do the fol- lowing:
Refer to Figure 6-1.
5-6 SERVICING ELECTRONIC CIRCUITRY
To check the heater blanket, verify the conti- nuity of the heater coil.
For troubleshooting the electronic system, re- fer to Section 4.0 and the appropriate pictorial diagrams at the back of the manual. The elec- tronic system utilizes printed circuit boards with solid-state components. After a malfunc- tion is traced to a particular board, the rec- ommended procedure is to return it to the factory for repair.
To check the temperature sensor, refer to Section 3-4 and measure its resistance when instrument power is off (should be about 440 ohms) and when instrument power is on (should range from 800 to 1,000 ohms). See Table 3-1.
To remove the glass converter tube (see Figure 6-4):
Rosemount Analytical Inc. A Division of Emerson Process Management Routine Servicing 5-7

5-8 Routine Servicing Rosemount Analytical Inc. A Division of Emerson Process Management


Tampering with or unauthorized substitution of components may adversely affect safety of this product. Use only factory-approved components for repair.
To identify the configuration of an analyzer, locate the analyzer name-rating plate. The sales matrix identifier number appears on the analyzer name-rating plate.
Each analyzer is configured per the customer sales order. Below is the 951C sales matrix which lists the various configurations avail- able.

951C Process Chemiluminescence NOx Analyzer (19" Rack Mount) (951C)
02 User controlled sample flow (no sample restrictor included)
Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 6-1

a. Common Parts
In most situations involving a malfunction of a circuit board, it is more practical to replace the board than to attempt isolation and replace- ment of the individual component. The cost of test and replacement will exceed the cost of a rebuilt assembly from the factory.
Refer to Figure 6-1.
655519 Air Restrictor Fitting 657091 Capacitor Assembly 655166 Capillary, Bypass 655589 Capillary, Sample Hi 623719 Capillary, Sample Lo 654068 Temperature Control Assembly 654070 Converter Assembly 655303 Exhaust Fan 654052 Fan Assembly 898587 Fuse 3.15 A 902413 Fuse 6.25 A 654390 I/O Assembly 652173 Ozone Generator 658156 Ozone Generator UV Lamp Replacement Kit 655129 Ozone Generator Power Supply 654062 Photomultiplier Assembly 655332 Power Supply Assembly 662273 Pressure Switch 623936 Sample Flow Restrictor 644055 Sample Pressure Gauge 815187 Sample Regulator 622917 Sensor, Temperature 6A00337G01/02/03 Signal/Control Board 654878 Transformer/Inductor Assembly
The following list does not include individual electronic components. If circumstances ne- cessitate replacement of an individual compo- nent which can be identified by inspection or from the schematic diagrams, obtain the re- placement component from a local source of supply.
The following parts are recommended for rou- tine maintenance and troubleshooting of the Model 951C NOX Analyzer. If the trouble- shooting procedures do not resolve the prob- lem, contact your local Rosemount Analytical service office. A list of Rosemount Analytical Service Centers is located in Section 7.
Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 6-3

Photomultiplier Assembly 654062 (See Figure 6-2)

Case Heater Temperature Control Assembly 654068 (See Figure 6-4)
Sample Regulator 815187 Sample Pressure Gauge 644055
Figure 6-1. Major Assemblies of the Model 951C
Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 6-3

b. Photomultiplier Assembly 654062
001522 O-Ring 008423 O-Ring, Photomultiplier 655168 Photomultiplier Tube 654381 Reaction Chamber 654086 Socket Assembly 639722 Thermal Shield
Refer to Figure 6-2.
654943 Housing 649541 Insulating Washer 636318 Magnetic Shield 630916 Magnetic Shield

Photomultiplier Tube and Magnetic Shield (PN 630916) to be Flush
Figure 6-2. Photomultiplier Housing Assembly
6-4 Replacement Parts Rosemount Analytical Inc. A Division of Emerson Process Management

c. Converter Assembly 654070
Refer to Figure 6-3.
632784 Connector, Blue Silicone 657127 Heater 632782 Temperature Sensor 632795 Tube

Figure 6-3. Converter Assembly
Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 6-5

d. Temperature Control Assembly 654068
Refer to Figure 6-4.
622733 Fan 622732 Heater 655335 Temperature Control Board 900492 Thermal Fuse

Figure 6-4. Case Heater Temperature Control Assembly
6-6 Replacement Parts Rosemount Analytical Inc. A Division of Emerson Process Management

7-1 RETURN OF MATERIAL If factory repair of defective equipment is re- quired, proceed as follows: 1. Secure a return authorization from a Rosemount Analytical Inc. Sales Office or Representative before returning the equipment. Equipment must be returned with complete identification in accordance with Rosemount instructions or it will not be accepted. Rosemount CSC (Customer Service Cen- ter) will provide the shipping address for your instrument. In no event will Rosemount be responsi- ble for equipment returned without proper authorization and identification. 2. Carefully pack the defective unit in a sturdy box with sufficient shock absorbing material to ensure no additional damage occurs during shipping. 3. In a cover letter, describe completely:
If warranty service is expected, the defective unit will be carefully inspected and tested at the factory. If the failure was due to the condi- tions listed in the standard Rosemount war- ranty, the defective unit will be repaired or replaced at Rosemount’s option, and an oper- ating unit will be returned to the customer in accordance with the shipping instructions fur- nished in the cover letter. For equipment no longer under warranty, the equipment will be repaired at the factory and returned as directed by the purchase order and shipping instructions.
7-2 CUSTOMER SERVICE For order administration, replacement Parts, application assistance, on-site or factory re- pair, service or maintenance contract informa- tion, contact:
Rosemount Analytical Inc.
7-3 TRAINING A comprehensive Factory Training Program of operator and service classes is available. For a copy of the Current Operator and Service Training Schedule contact :
Rosemount Analytical Inc.
Rosemount Analytical Inc.
7-2 Replacement Parts Rosemount Analytical Inc. A Division of Emerson Process Management

7-2 Return of Material Rosemount Analytical Inc. A Division of Emerson Process Management

INDEX
establish correct pressure of zero gas, 3-4 Excessive Photomultiplier Dark Current, 5-2
F flow rate, measuring, 5-3 flow rate, subnormal, 5-3
A Air (U.S.P. Breathing Grade), 2-5 AIR inlet
adequate air pressure, 3-3 setpoint pressure, 3-3 analyzer
I Inability to obtain a flow of one liter per minute at the
EXHAUST outlet, 3-4 internal SAMPLE pressure, 3-1
installation location, 2-3 mount near sample, 2-3 operating temperature, 2-3 warm-up, 7, 3-3
C Contamination of Reaction Chamber or Sample Flow
routing, 2-3 ozonator air
supply pressure, 7 ozone generation system, description, 5-6 ozone generation system, power supply removal, 5-
glass converter tube, 5-7 heater blanket, 5-7 measuring efficiency, 3-9 optimizing operating temperature, 3-7 temperature check, 3-7 temperature sensor, 5-7 converter temperature
7 ozone generation system, removal, 5-6 ozone generation system, UV lamp replacement, 5-7 ozone lamp
output check, 5-2 OZONE pressure
adjustment, 3-7 recommended periodic check, 3-7 current output
determination, 3-1 nominal, 3-1
adjustment, 2-3 Current recorder connections
P photomultiplier tube, 1-1, 4-1, 5-1, 5-2, 5-3, 5-4, 5-6 photomultiplier tube, description, 5-5 potentiometric recorder connections
cable specifications, 2-3 hookup, 2-3 routing, 2-3 cylinders, air/gas
cable specifications, 2-3 hookup, 2-3 routing, 2-3 Power connections
cable type, 2-3 hookup, 2-2 routing, 2-3 PPM RANGE Switch, 3-3
E Establish correct flow of upscale standard gas, 3-4 Establish correct pressure for air, 3-4 Establish correct pressure of sample gas, 3-4
Rosemount Analytical Inc. A Division of Emerson Process Management Index 8-1
adjustment, 3-1 · System, 5-2 converter · location, 2-3

reaction chamber, 4-1, 4-2, 5-1, 5-2, 5-3, 5-4, 5-5, 5-
6 reaction chamber, alternate cleaning procedure, 5-5 reaction chamber, standard cleaning procedure, 5-5 Return Authorization, 7-2
PPM RANGE switch, 3-6 RANGE1 control, 3-6 sensitivity, 3-6 supply gas, 3-6
Z Zero Calibration, 7
establish correct pressure, 7 supply pressure, 7 shutdown, 3-6 Span Gas, 2-5 supply pressure, 2-5
PPM RANGE switch, 3-6 RANGE1 control, 3-6 ZERO control, 3-6 zero display reading, inability to obtain, 5-2 zero gas
establish correct pressure, 7 supply pressure, 7
8-4 Index Rosemount Analytical Inc. A Division of Emerson Process Management


Drawing 1-1. Panel Cutout / Installation Drawing
Rosemount Analytical Inc. A Division of Emerson Process Management List of Drawings


Drawing 1-2 Flow Diagram, Lo/Mid Range
Rosemount Analytical Inc. A Division of Emerson Process Management
List of Drawings


Drawing 1-3 Flow Diagram, Hi Range
Rosemount Analytical Inc. A Division of Emerson Process Management List of Drawings

Goods and part(s) (excluding consumables) manufactured by Seller are warranted to be free from defects in workmanship and material under normal use and service for a period of twelve (12) months from the date of shipment by Seller. Consumables, glass electrodes, membranes, liquid junctions, electrolyte, o-rings, etc., are warranted to be free from defects in workmanship and material under normal use and service for a period of ninety (90) days from date of shipment by Seller. Goods, part(s) and consumables proven by Seller to be defective in workmanship and/or material shall be replaced or repaired, free of charge, F.O.B. Seller's factory provided that the goods, part(s) or consumables are returned to Seller's designated factory, transportation charges prepaid, within the twelve (12) month period of warranty in the case of goods and part(s), and in the case of consumables, within the ninety (90) day period of warranty. This warranty shall be in effect for replacement or repaired goods, part(s) and the remaining portion of the ninety (90) day warranty in the case of consumables. A defect in goods, part(s) and consumables of the commercial unit shall not operate to condemn such commercial unit when such goods, part(s) and consumables are capable of being renewed, repaired or replaced.
The Seller shall not be liable to the Buyer, or to any other person, for the loss or damage directly or indirectly, arising from the use of the equipment or goods, from breach of any warranty, or from any other cause. All other warranties, expressed or implied are hereby excluded.
IN CONSIDERATION OF THE HEREIN STATED PURCHASE PRICE OF THE GOODS, SELLER GRANTS ONLY THE ABOVE STATED EXPRESS WARRANTY. NO OTHER WARRANTIES ARE GRANTED INCLUDING, BUT NOT LIMITED TO, EXPRESS AND IMPLIED WARRANTIES OR MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
Limitations of Remedy. SELLER SHALL NOT BE LIABLE FOR DAMAGES CAUSED BY DELAY IN PERFORMANCE. THE SOLE AND EXCLUSIVE REMEDY FOR BREACH OF WARRANTY SHALL BE LIMITED TO REPAIR OR REPLACEMENT UNDER THE STANDARD WARRANTY CLAUSE. IN NO CASE, REGARDLESS OF THE FORM OF THE CAUSE OF ACTION, SHALL SELLER'S LIABILITY EXCEED THE PRICE TO BUYER OF THE SPECIFIC GOODS MANUFACTURED BY SELLER GIVING RISE TO THE CAUSE OF ACTION. BUYER AGREES THAT IN NO EVENT SHALL SELLER'S LIABILITY EXTEND TO INCLUDE INCIDENTAL OR CONSEQUENTIAL DAMAGES. CONSEQUENTIAL DAMAGES SHALL INCLUDE, BUT ARE NOT LIMITED TO, LOSS OF ANTICIPATED PROFITS, LOSS OF USE, LOSS OF REVENUE, COST OF CAPITAL AND DAMAGE OR LOSS OF OTHER PROPERTY OR EQUIPMENT. IN NO EVENT SHALL SELLER BE OBLIGATED TO INDEMNIFY BUYER IN ANY MANNER NOR SHALL SELLER BE LIABLE FOR PROPERTY DAMAGE AND/OR THIRD PARTY CLAIMS COVERED BY UMBRELLA INSURANCE AND/OR INDEMNITY COVERAGE PROVIDED TO BUYER, ITS ASSIGNS, AND EACH SUCCESSOR INTEREST TO THE GOODS PROVIDED HEREUNDER.
Force Majeure. Seller shall not be liable for failure to perform due to labor strikes or acts beyond Seller's direct control.
Rosemount Analytical Inc. A Division of Emerson Process Management



ROSEMOUNT ANALYTICAL EUROPE Emerson Process Management GmbH & Co. OHG
Emerson Process Management Rosemount Analytical
Emerson Process Management Rosemount Analytical Inc.
Industriestrasse 1 63594 Hasselroth
11100 Brittmoore Park Road Houston, TX 77041 USA T 713.467.6000 F 713.827.3328F 49 6055 884209
6565 P Davis Industrial Parkway Solon, OH 44139 USA
Germany T 49 6055 884 0 F 49 6055 884209
T 440.914.1261 Toll Free in US and Canada 800.433.6076 F 440.914.1271
ASIA-PACIFIC Emerson Process Management
EUROPE Emerson Process Management Shared Services Limited
MIDDLE EAST AND AFRICA Emerson Process Management
Asia Pacific Private Limited
EPM Building P.O. Box 17033 Jebel Ali Free Zone
1 Pandan Crescent Singapore 128461
Heath Place Bognor Regis
Republic of Singapore T 65 6 777 8211
West Sussex PO22 9SH England
Dubai, United Arab Emirates T 971 4 8835235 F 971 4 8835312
F 65 6 777 0947 e-mail: analytical@ap.emersonprocess.com
© Rosemount Analytical Inc. 2009