Emerson Controller (951C)

64 pages·1.8 MB·Controller·PDF·ENDEFRITNLJA
Emerson Controller (951C)
Table of Contents 80 sections Model 951C NO X Analyzer · ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING! · HIGHLIGHTS OF CHANGES · EFFECTIVE JUNE 10, 2009 · … ShowHide
Model 951C NO X Analyzerp.1ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING!p.2HIGHLIGHTS OF CHANGESp.3EFFECTIVE JUNE 10, 2009p.3PAGE SUMMARYp.3TABLE OF CONTENTSp.4LIST OF ILLUSTRATIONSp.6LIST OF TABLESp.6LIST OF DRAWINGSp.6PREFACEp.9DEFINITIONSp.9NOTEp.9SAFETY SUMMARYp.10AUTHORIZED PERSONNELp.10ELECTRICAL SHOCK HAZARDp.10WARNINGp.10INTERNAL ULTRAVIOLET LIGHT HAZARDp.10TOXIC CHEMICAL HAZARDp.11PARTS INTEGRITYp.11HIGH PRESSURE GAS CYLINDERSp.11TOXIC AND OXIDIZING GAS HAZARDp.12HANDLE LAMP ASSEMBLY WITH EXTREME CAREp.12TOPPLING HAZARDp.12GENERAL PRECAUTIONS FOR HANDLING AND STORING HIGHp.13PRESSURE GAS CYLINDERSp.13DOCUMENTATIONp.14CONDENSED STARTUP AND CALIBRATION PROCEDUREp.15SECTION 1 DESCRIPTION AND SPECIFICATIONSp.171-1 OVERVIEWp.171-2 TYPICAL APPLICATIONSp.171-3 SPECIFICATIONSp.18SECTION 2 INSTALLATIONp.192-1 UNPACKINGp.192-2 LOCATIONp.192-4 ELECTRICAL CONNECTIONSp.192-3 VOLTAGE REQUIREMENTSp.19S3p.20INTERIOR EXTERIORp.21RECORDERp.22POWERp.22OUTPUTp.22J2p.22RANGE 1p.22RANGE 2p.22RANGE 3p.22RANGE 4p.22RTNp.22+24 VDC INp.22CHS GNDp.22TB1p.22L1/HOTp.22FUSEp.22L2/NEUTp.22GNDp.22CUR OUTPUTp.22VOLT OUTPUTp.22+ - G + -p.22TB2p.222-5 GAS REQUIREMENTSp.232-6 SAMPLE REQUIREMENTSp.232-7 GAS CONNECTIONSp.23TOXIC AND OXIDIZING GAS HAZARDSp.232-8 LEAK TESTp.24SECTION 3 OPERATIONp.253-1 FRONT PANEL INDICATORS AND CONTROLSp.25R25 R43p.263-2 STARTUP PROCEDUREp.273-3 CALIBRATIONp.303-4 ROUTINE OPERATIONp.303-5 CONVERTER TEMPERATURE ADJUSTMENT PROCEDUREp.30TEMPERATUREp.31RESISTANCEp.31( ° C)p.313-6 MEASUREMENT OF CONVERTER EFFICIENCYp.333-7 RECOMMENDED CALIBRATION FREQUENCYp.33SECTION 4p.35THEORYp.354-1 NITRIC OXIDE DETERMINATION BY CHEMILUMINESCENCE METHODp.354-2 ANALYZER FLOW SYSTEMp.35UV 3O 2 → 2O 3p.35
Page 1 of the Emerson Controller (951C) user manual — Model 951C NO X Analyzer
Page 1 · click to enlarge
Page 1

Model 951C NO X Analyzer

Emerson Controller (951C) — Model 951C NO X Analyzer diagram (page 1)
Fig. 1 · Model 951C NO X Analyzer
Emerson Controller (951C) — Model 951C NO X Analyzer diagram (page 1)
Fig. 2 · Model 951C NO X Analyzer
Emerson Controller (951C) — Model 951C NO X Analyzer diagram (page 1)
Fig. 3 · Model 951C NO X Analyzer

http://www.raihome.com

Page 2 of the Emerson Controller (951C) user manual — ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING!
Page 2 · click to enlarge
Page 2

ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING!

Rosemount Analytical designs, manufactures and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you MUST properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions MUST be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount Analytical products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.

  • Read all instructions prior to installing, operating, and servicing the product.
  • If you do not understand any of the instructions, contact your Rosemount Analytical representative for clarification.
  • Follow all warnings, cautions, and instructions marked on and supplied with the product.
  • Inform and educate your personnel in the proper installation, operation, and maintenance of the product .
  • Install your equipment as specified in the Installation Instructions of the appropriate Instruction Manual and per applicable local and national codes . Connect all products to the proper electrical and pressure sources.
  • To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the product.
  • When replacement parts are required, ensure that qualified people use replacement parts specified by Rosemount. Unauthorized parts and procedures can affect the product’s performance, place the safe operation of your process at risk, and VOID YOUR WARRANTY . Look-alike substitutions may result in fire, electrical hazards, or improper operation.
  • Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being performed by qualified persons, to prevent electrical shock and personal injury .

The information contained in this document is subject to change without notice.

Teflon® is a registered trademark of E.I. duPont de Nemours and Co., Inc. Alconox is a registered trademark of Alconox, Inc. SNOOP® is a registered trademark of NUPRO Co.

WORLD HEADQUARTERS Emerson Process Management Rosemount Analytical Inc. 6565P Davis Industrial Parkway Solon, OH 44139 T 440 914 1261 T 800 433 6076 F 440 914 1262 E gas.csc@emerson.com

Page 3 of the Emerson Controller (951C) user manual — EFFECTIVE JUNE 10, 2009
Page 3 · click to enlarge
Page 3

HIGHLIGHTS OF CHANGES

EFFECTIVE JUNE 10, 2009

PAGE SUMMARY

Cover Updated photo, revision number, and release date

P-6 Removed Compliance section

1-1 Added Mid Range

1-2 Added Mid Range

3-1 Added Mid Range, Backlight over range feature, and a note, using the MID range, set the pressure regulator to 2.5 psig.

3-2 Added a photo of the Signal Board, Added a photo of the Signal Board Test Points, removed the signal board diagnostic jumpers (Figure 3-4)

3-3 Removed the Hi/Lo Range Select Jumper (Figure 3-5), Updated the PPM Range selection switch photo and table, removed the decimal point selection jumpers (Figure 3-7), added Configuration Jumper Settings (Figure 3-7)

3-6 Added step 3 to the Zero Calibration Section (In the Mid Range Configuration, Set the Sample Pressure Regulator to 2.5 psig), Added step 3 to the span cali- bration section (In the Mid Range Configuration, Set the Sample Pressure Regu- lator to 2.5 psig), Added step 6 to the span calibration section (Use R25 and R43 for fine adjustments to the display and recorder output.)

4-3 Updated the Signal Conditioning and Display section to reflect the current PCB’s, updated the Circuit function section.

4-5 Updated the Analyzer Signal Conditioning Block Diagram

6-1 Updated the MATRIX

6-3 Updated the Replacement Parts List

List of Drawings Added Drawing 1-1, 1-2, and1-3

Back Cover Updated Company addresses and contact information

Rosemount Analytical Inc. A Division of Emerson Process Management Contents i

Page 4 of the Emerson Controller (951C) user manual — Table of Contents
Page 4 · Contents
Contents

Contents

Model 951C NO X Analyzerp.1ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING!p.2HIGHLIGHTS OF CHANGESp.3EFFECTIVE JUNE 10, 2009p.3PAGE SUMMARYp.3TABLE OF CONTENTSp.4LIST OF ILLUSTRATIONSp.6LIST OF TABLESp.6LIST OF DRAWINGSp.6PREFACEp.9DEFINITIONSp.9NOTEp.9SAFETY SUMMARYp.10AUTHORIZED PERSONNELp.10ELECTRICAL SHOCK HAZARDp.10WARNINGp.10INTERNAL ULTRAVIOLET LIGHT HAZARDp.10TOXIC CHEMICAL HAZARDp.11PARTS INTEGRITYp.11HIGH PRESSURE GAS CYLINDERSp.11TOXIC AND OXIDIZING GAS HAZARDp.12HANDLE LAMP ASSEMBLY WITH EXTREME CAREp.12TOPPLING HAZARDp.12GENERAL PRECAUTIONS FOR HANDLING AND STORING HIGHp.13PRESSURE GAS CYLINDERSp.13DOCUMENTATIONp.14CONDENSED STARTUP AND CALIBRATION PROCEDUREp.15SECTION 1 DESCRIPTION AND SPECIFICATIONSp.171-1 OVERVIEWp.171-2 TYPICAL APPLICATIONSp.171-3 SPECIFICATIONSp.18SECTION 2 INSTALLATIONp.192-1 UNPACKINGp.192-2 LOCATIONp.192-4 ELECTRICAL CONNECTIONSp.192-3 VOLTAGE REQUIREMENTSp.19S3p.20INTERIOR EXTERIORp.21RECORDERp.22POWERp.22OUTPUTp.22J2p.22RANGE 1p.22RANGE 2p.22RANGE 3p.22RANGE 4p.22RTNp.22+24 VDC INp.22CHS GNDp.22TB1p.22L1/HOTp.22FUSEp.22L2/NEUTp.22GNDp.22CUR OUTPUTp.22VOLT OUTPUTp.22+ - G + -p.22TB2p.222-5 GAS REQUIREMENTSp.232-6 SAMPLE REQUIREMENTSp.232-7 GAS CONNECTIONSp.23TOXIC AND OXIDIZING GAS HAZARDSp.232-8 LEAK TESTp.24SECTION 3 OPERATIONp.253-1 FRONT PANEL INDICATORS AND CONTROLSp.25R25 R43p.263-2 STARTUP PROCEDUREp.273-3 CALIBRATIONp.303-4 ROUTINE OPERATIONp.303-5 CONVERTER TEMPERATURE ADJUSTMENT PROCEDUREp.30TEMPERATUREp.31RESISTANCEp.31( ° C)p.313-6 MEASUREMENT OF CONVERTER EFFICIENCYp.333-7 RECOMMENDED CALIBRATION FREQUENCYp.33SECTION 4p.35THEORYp.354-1 NITRIC OXIDE DETERMINATION BY CHEMILUMINESCENCE METHODp.354-2 ANALYZER FLOW SYSTEMp.35UV 3O 2 → 2O 3p.35
Page 6 of the Emerson Controller (951C) user manual — LIST OF DRAWINGS
Page 6 · click to enlarge
Page 6

LIST OF ILLUSTRATIONS

Figure 2-1. Power Supply Board Voltage Select Switches 2-2 Figure 2-2. Temperature Control Board 2-3 Figure 2-3. Cable Gland 2-3 Figure 2-4. Rear View of Model 951C (cover removed) 2-4 Figure 2-5. Rear Panel Wiring Connections 2-4 Figure 3-1. Model 951C Controls, Indicators and Adjustments 3-2 Figure 3-2. Signal Board 3-3 Figure 3-3. Signal Board Test Points 3-3 Figure 3-6. PPM Range Select Switch 3-3 Figure 3-7. Configuration Jumper Settings 3-3 Figure 3-8. Range 2 Selection Jumpers 3-3 Figure 3-9. Power Supply Board 3-5 Figure 4-1. Analyzer Block Diagram 4-4 Figure 4-2. Analyzer Thermal System 4-5 Figure 6-1. Major Assemblies of the Model 951C 6-3 Figure 6-2. Photomultiplier Housing Assembly 6-4 Figure 6-3. Converter Assembly 6-5 Figure 6-4. Case Heater Temperature Control Assembly 6-6

LIST OF TABLES

Table 3-1. Resistance of Converter Temperature Sensor vs. Temperature 3-7

LIST OF DRAWINGS

Drawing 1-1 Panel Cutout / Installation Drawing Drawing 1-2 Flow Diagram, Lo/Mid Range Drawing 1-3 Flow Diagram, Hi Range

iv Contents Rosemount Analytical Inc. A Division of Emerson Process Management

Page 7 of the Emerson Controller (951C) user manual — LIST OF ILLUSTRATIONS
Page 7 · click to enlarge
Page 7

Rosemount Analytical Inc. A Division of Emerson Process Management Contents v

Page 8 of the Emerson Controller (951C) user manual — LIST OF ILLUSTRATIONS
Page 8 · click to enlarge
Page 8
Page 8 of the Emerson Controller (951C) user manual — LIST OF ILLUSTRATIONS
Page 9 of the Emerson Controller (951C) user manual — NOTE
Page 9 · click to enlarge
Page 9

PREFACE

The purpose of this manual is to provide information concerning the components, functions, installation and maintenance of the 951C NOX Analyzer.

Some sections may describe equipment not used in your configuration. The user should become thoroughly familiar with the operation of this module before operating it. Read this instruction manual completely.

DEFINITIONS

The following definitions apply to DANGERS, WARNINGS, CAUTIONS and NOTES found throughout this publication.

Highlights the presence of a hazard which will cause severe personal injury, death, or substantial property damage if the warning is ignored.

Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in injury, death, or long-term health hazards of personnel.

Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in damage to or destruction of equipment, or loss of effectiveness.

NOTE

Highlights an essential operating procedure, condition or statement.

Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-1

Page 10 of the Emerson Controller (951C) user manual — WARNING
Page 10 · click to enlarge
Page 10

SAFETY SUMMARY

If this equipment is used in a manner not specified in these instructions, protective systems may be impaired.

AUTHORIZED PERSONNEL

To avoid explosion, loss of life, personal injury and damage to this equipment and on-site property, all personnel authorized to install, operate and service the this equipment should be thoroughly familiar with and strictly follow the instructions in this manual. SAVE THESE IN- STRUCTIONS.

ELECTRICAL SHOCK HAZARD

Do not operate without doors and covers secure. Servicing requires access to live parts which can cause death or serious injury. Refer servicing to qualified personnel.

This instrument was shipped from factory set up to operate on 115 volt 50/60 Hz. For operation on 230 volt 50/60 Hz, refer to Section 2-3.

For safety and proper performance this instrument must be connected to a properly grounded three-wire source of power.

WARNING

INTERNAL ULTRAVIOLET LIGHT HAZARD

Ultraviolet light from the ozone generator can cause permanent eye damage. Do not look directly at the ultraviolet source in ozone generator. Use of ultraviolet filtering glasses is recommended.

P-2 Preface Rosemount Analytical Inc. A Division of Emerson Process Management

Page 11 of the Emerson Controller (951C) user manual — HIGH PRESSURE GAS CYLINDERS
Page 11 · click to enlarge
Page 11

WARNING

TOXIC CHEMICAL HAZARD

This instrument generates ozone which is toxic by inhalation and is a strong irritant to throat and lungs. Ozone is also a strong oxidizing agent. Its presence is detected by a characteristic pungent odor.

The instrument exhaust contains both ozone and nitrogen dioxide, both toxic by inhalation, and may contain other constituents of the sample gas which may be toxic. Such gases include various oxides of nitrogen, unburned hydrocarbons, carbon monoxide and other products of combustion reactions. Carbon monoxide is highly toxic and can cause headache, nausea, loss of consciousness, and death.

Avoid inhalation of the ozone produced within the analyzer and avoid inhalation of the sample and exhaust products transported within the analyzer. Avoid inhalation of the combined exhaust products at the exhaust fitting.

Keep all tube fittings tight to avoid leaks. See Section 2-8 for Leak Test Procedure.

Connect rear exhaust outlet to outside vent by a 1/4 inch (6.3 mm) or larger stainless steel or Teflon line. Check vent line and connections for leakage.

PARTS INTEGRITY

Tampering or unauthorized substitution of components may adversely affect safety of this product. Use only factory documented components for repair.

HIGH PRESSURE GAS CYLINDERS

This instrument requires periodic calibration with a known standard gas. See Sections 2-5 and 3-3. See also General Precautions for Handling and Storing High Pressure Gas Cylinders, page P-5.

Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-3

Page 12 of the Emerson Controller (951C) user manual — TOPPLING HAZARD
Page 12 · click to enlarge
Page 12

TOXIC AND OXIDIZING GAS HAZARD

The ozone generator lamp contains mercury. Lamp breakage could result in mercury exposure. Mercury is highly toxic if absorbed through skin or ingested, or if vapors are inhaled.

HANDLE LAMP ASSEMBLY WITH EXTREME CARE

If lamp is broken, avoid skin contact and inhalation in the area of the lamp or the mercury spill.

  • Immediately clean up and dispose of the mercury spill and lamp residue as follows:
  • Wearing rubber gloves and goggles, collect all droplets of mercury by means of a suction pump and aspirator bottle with long capillary tube. Alternatively, a commercially available mercury spill clean-up kit, such as J. T. Baker product No. 4439-01, is recommended.
  • Carefully sweep any remaining mercury and lamp debris into a dust pan. Carefully transfer all mercury, lamp residue and debris into a plastic bottle which can be tightly capped. Label and return to hazardous material reclamation center.
  • Do not place in trash, incinerate or flush down sewer.
  • Cover any fine droplets of mercury in non-accessible crevices with calcium polysulfide and sulfur dust.

TOPPLING HAZARD

This instrument’s internal pullout chassis is equipped with a safety stop latch located on the left side of the chassis.

When extracting the chassis, verify that the safety latch is in its proper (counter-clockwise) orientation.

If access to the rear of the chassis is required, the safety stop may be overridden by lifting the latch; however, further extraction must be done very carefully to insure the chassis does not fall out of its enclosure.

If the instrument is located on top of a table or bench near the edge, and the chassis is extracted, it must be supported to prevent toppling.

Failure to observe these precautions could result in personal injury and/or damage to the product.

P-4 Preface Rosemount Analytical Inc. A Division of Emerson Process Management

Page 13 of the Emerson Controller (951C) user manual — GENERAL PRECAUTIONS FOR HANDLING AND STORING HIGH
Page 13 · click to enlarge
Page 13

GENERAL PRECAUTIONS FOR HANDLING AND STORING HIGH

PRESSURE GAS CYLINDERS

Edited from selected paragraphs of the Compressed Gas Association's "Handbook of Compressed Gases" published in 1981

Compressed Gas Association 1235 Jefferson Davis Highway Arlington, Virginia 22202

1. Never drop cylinders or permit them to strike each other violently.
2. Cylinders may be stored in the open, but in such cases, should be protected against extremes of weather and, to prevent rusting, from the dampness of the ground. Cylinders should be stored in the shade when located in areas where extreme temperatures are prevalent.
3. The valve protection cap should be left on each cylinder until it has been secured against a wall or bench, or placed in a cylinder stand, and is ready to be used.
4. Avoid dragging, rolling, or sliding cylinders, even for a short distance; they should be moved by using a suitable hand-truck.
5. Never tamper with safety devices in valves or cylinders.
6. Do not store full and empty cylinders together. Serious suckback can occur when an empty cylinder is attached to a pressurized system.
7. No part of cylinder should be subjected to a temperature higher than 125 ° F (52 ° C). A flame should never be permitted to come in contact with any part of a compressed gas cylinder.
8. Do not place cylinders where they may become part of an electric circuit. When electric arc welding, precautions must be taken to prevent striking an arc against the cylinder.

Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-5

Used by Permission

Page 14 of the Emerson Controller (951C) user manual — DOCUMENTATION
Page 14 · click to enlarge
Page 14

DOCUMENTATION

The following 951C instruction materials are available. Contact Customer Service Center or the local rep- resentative to order.

748214 Instruction Manual (this document)

P-6 Preface Rosemount Analytical Inc. A Division of Emerson Process Management

Page 15 of the Emerson Controller (951C) user manual — NOTE
Page 15 · click to enlarge
Page 15

CONDENSED STARTUP AND CALIBRATION PROCEDURE

The following summarized instructions on startup and calibration are intended for opera- tors already familiar with the analyzer.

8. Establish correct pressure of zero gas:

a. Supply zero gas to rear panel SAM-

PLE inlet and set to 15 psig.

For initial startup, refer to detailed instructions provided in Section 3.

b. Note reading on SAMPLE Pressure

Gauge. It should be the same as in Step 6b. If not, adjust output pres- sure regulator on the zero gas cylin- der as required.

1. Review the Purchase Order from the 951C Analyzer. Make a note of the Range that was purchase (Low Range, Mid Range, or High Range).
7. Establish correct pressure of upscale standard gas:
2. Set Range switch on the Signal Conditioning Board to position 4, 250ppm, 500ppm, or 2500 ppm (see Figure 3-2, page 3-2).

a. Supply upscale standard gas to rear

panel SAMPLE inlet.

3. On the Signal Conditioning Board, Verify that Hi/Mid/Lo Range Configuration Jumpers are installed for the Range that was purchase (Figure 3-7, page 3-3)

b. Note reading on SAMPLE Pressure

Gauge. It should be the same as in Step 7b. If not, adjust output regula- tor on cylinder of upscale standard gas as required.

4. Apply power to the analyzer. The analyzer will now require approximately one to two hours for temperature equilibrium before being ready for calibration.

NOTE

Supply pressure for sample, upscale standard gas and zero air must be the same. If not, the readout will be in error.

5. Verify that the pressure regulator on the cylinder of zero gas (nitrogen or air) or sample gas is set for supply pressure of 10 to 17 psig.
8. Zero Calibration:

a. Set PPM RANGE Switch (Figure

6. Verify that the pressure regulator on the cylinder of air (ozonator supply) is set for supply pressure of 20 to 25 psig.

3-6, page 3-3) for range to be used for sample analysis. Set the front panel RANGE1 Control at normal operating setting, if known, or at about middle or the range if normal setting is not known.

7. Establish correct pressure of sample gas:

a. Supply sample gas to rear-panel SAM-

PLE inlet at 10 to 17 psig (normally 15 psig).

b. Supply zero gas to rear panel SAM-

PLE inlet.

b. Adjust SAMPLE Back Pressure Regula-

c. Adjust front panel ZERO Control for

reading of zero on meter or recorder.

tor so that SAMPLE Pressure Gauge in- dicates the value appropriate to the desired operating range (normal operat- ing pressure is 3 to 5 psig). Using the Mid range setting, set the sample pres- sure regulator to 2.5 psig See Figure 3-1.

9. Upscale Calibration:

a. Set PPM RANGE Switch (Figure

3-6, page 3-3) at setting appropriate to the particular span gas.

Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-7

Page 16 of the Emerson Controller (951C) user manual — NOTE
Page 16 · click to enlarge
Page 16

b. Supply upscale standard gas of accu-

NOTE

rately known NOX content to rear panel SAMPLE inlet.

It is the responsibility of the user to measure efficiency of the NO2-to-NO converter during initial startup, and thereafter at intervals appropriate to the application, normally once a month.

c. Adjust front panel RANGE1 Control so

that reading on meter or recorder is equal to the know parts-per-million con- centration of NOX in the span gas.

d. Adjust R25 on the signal board so that

the Display value and the recorder out- put are equal.

P-8 Preface Rosemount Analytical Inc. A Division of Emerson Process Management

Page 17 of the Emerson Controller (951C) user manual — 1-2 TYPICAL APPLICATIONS
Page 17 · click to enlarge
Page 17

SECTION 1 DESCRIPTION AND SPECIFICATIONS

1-1 OVERVIEW

The case heater assembly of the Model 951C maintains the internal temperature at approximately 50oC (122oF).

The Model 951C NOX Analyzer is designed to measure NOX using one of three sets of ranges designated as Hi, Mid, or Lo. The Hi Range set consists of spans with ranges of 0-100, 0-250, 0-1000, and 0-2500 ppm NOX. The Mid Range set consists of spans with ranges of 0-20, 0-50, 0-200, and 0-500 ppm NOX. The Lo Range set consists of spans with ranges of 0-10, 0-25, 0-100, and 0-250 ppm NOX.

1-2 TYPICAL APPLICATIONS

The Model 951C Analyzer has specific ap- plications in the following areas:

  • Oxides of nitrogen (NO X ) emissions from the combustion of fossil fuels in:

The NOX analyzer continuously analyzes a flow- ing gas sample for NOX [nitric oxide (NO) plus nitrogen dioxide (NO2)]. The sum of the concen- trations is continuously reported as NOX.

  • Vehicle engine exhaust
  • Incinerators

The analyzer is based on the chemilumines- cence method of NO detection. The sample is continuously passed through a heated bed of vitreous carbon, in which NO2 is reduced to NO. Any NO initially present in the sample passes through the converter unchanged, and any NO2 is converted to an approximately equivalent (95%) amount of NO.

  • Boilers
  • Gas appliances
  • Turbine exhaust
  • Nitric acid plant emissions
  • Ammonia in pollution control equipment (with converter)

The NO is quantitatively converted to NO2 by gas-phase oxidation with molecular ozone pro- duced within the analyzer from air supplied by an external cylinder. During this reaction, ap- proximately 10% of the NO2 molecules are ele- vated to an electronically excited state, followed by immediate decay to the non-excited state, accompanied by emission of photons. These photons are detected by a photomultiplier tube, which in turn generates a DC current propor- tional to the concentration of NOX in the sample stream. The current is then amplified and used to drive a front panel display and to provide po- tentiometric and isolated current outputs.

  • Nitric oxide emissions from decaying organic material (i.e., landfills).

To minimize system response time, an internal sample-bypass feature provides high-velocity sample flow through the analyzer.

The Backlight blinks when the analyzer is 10% or more over-range. Selecting a less sensitive (higher) range restores the display function.

Rosemount Analytical Inc. A Division of Emerson Process Management Installation 1-1

Page 18 of the Emerson Controller (951C) user manual — 1-3 SPECIFICATIONS
Page 18 · click to enlarge
Page 18

1-3 SPECIFICATIONS

LO 0 to 10, 0 to 25, 0 to 100, 0 to 250 ppm NOX MID 0 to 20, 0 to 50, 0 to 200, 0 to 500 ppm NOX HI 0 to 100, 0 to 250, 0 to 1000, 0 to 2500 ppm NOX Repeatability within 0.1 ppm or ±1% of fullscale, whichever is greater Zero/Span Drift less than ±0.1 ppm or ±1% of fullscale, whichever is greater, in 24

hours at constant temperature less than ±0.2 ppm or ±2% of fullscale, whichever is greater, over any 10°C interval from 4 to 40°C (for rate change of 10°C or less per hour) Response Time

(Electronic + Flow) 90% of fullscale in less than 1 minute

Sensitivity less than 0.1 ppm or 1% of fullscale, whichever is greater Detector Operating Pressure atmospheric Total Sample Flow Rate 1 Liter per minute at 20 psig Sample Pressure 138 kPa (20 psig) Ozone Generator Gas U.S.P. breathing-grade air

Ambient Temperature Range 4 to 40°C (40 to 104°F) Analog Output

Potentiometric 0 to +5 VDC, 2000 ohm minimum load Isolated Current Field-selectable 0 to 20 or 4 to 20 mA, 700 ohm max load Display Digital, 4-1/2 digit LCD, readout in engineering units, back-lighted Power Requirements 115/230 VAC ±10%, 50/60 ±3 Hz, 570 W maximum Enclosure General purpose for installation in weather-protected areas Dimensions 8.7 in. x 19.0 x 19.0 in. (H x W x D)

22.0 cm x 48.3 cm x 48.3 cm (H x W x D) Weight 22.2 kg (49 lbs) approximate

Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management 1-2

Ranges

Page 19 of the Emerson Controller (951C) user manual — 2-3 VOLTAGE REQUIREMENTS
Page 19 · click to enlarge
Page 19

SECTION 2 INSTALLATION

2-1 UNPACKING

the Power Supply Board (Figure 3-9, page 3-5) position voltage select switches S1, S2, S3 (Figure 2-1, page 2-2) and S3 (located on the Temperature Control Board (Figure 2-2, page 2- 2) in the 230 VAC position.

Carefully examine the shipping carton and con- tents for signs of damage. Immediately notify the shipping carrier if the carton or its contents are damaged. Retain the carton and packing material until the instrument is operational.

Refer to Figure 2-4, page 2-3. On the rear of the analyzer, remove the 6.25 A fuse (P/N 902413) and replace with the 3.15 A fuse (P/N 898587) provided in the shipping kit.

2-2 LOCATION

See drawing 654063 for Outline and Mounting dimensions.

2-4 ELECTRICAL CONNECTIONS

Install analyzer in a clean area, free from mois- ture and excessive vibration, at a stable tem- perature within 4 to 40°C.

The power and output (recorder and current) cable glands are supplied loose in the shipping kit to allow cable installation to connectors or terminal strips.

The analyzer should be mounted near the sam- ple source to minimize sample-transport time.

Cable Gland Part No. Power 899330 Recorder 899329

A temperature control system maintains the in- ternal temperature of analyzer at 50°C (122°F) to ensure proper operation over an ambient temperature range of 4°C to 40°C (40°F to 104°F). Temperatures outside these limits ne- cessitate use of special temperature-controlling equipment or environmental protection. Also, the ambient temperature should not change at a rate exceeding 10°C/hr.

Remove rear cover to access terminals. Route each cable through the cable gland and connect to the appropriate connector or terminal strip, tighten the gland.

a. Line Power Connections

Refer to Figure 2-3 (page 2-3), Figure 2-4 (page 2-3), Figure 2-5 (page 2-4) and draw- ing 654063. If this instrument is located on a bench or table top or is installed in a pro- tected rack, panel or cabinet, power may be connected via a 3-wire flexible power cord, minimum 18 AWG (max. O.D. 0.480", min. O./D. 0.270"), through the hole labeled POWER, utilizing connector gland (P/N 899330) provided.

The cylinders of air and span gas should be lo- cated in an area of constant ambient tempera- ture (±10°C).

2-3 VOLTAGE REQUIREMENTS

WARNING

ELECTRICAL SHOCK HAZARD

For safety and proper performance this instrument must be connected to a properly grounded three-wire source of power.

Route the power cable through the cable gland and connect the leads to TB1 on the rear panel. Tighten the cable gland ade- quately to prevent rotation or slippage of the power cable. Since the rear terminals do not slide out with the chassis, no excess power cable slack is necessary.

This instrument was shipped from the factory set up to operate on 115 VAC, 50/60 Hz electric power. For operation on 230 VAC, 50/60 Hz, on

Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-3

Page 20 of the Emerson Controller (951C) user manual — S3
Page 20 · click to enlarge
Page 20

The following power cord and/or support feet (for bench top use) are available:

  • North American power cord set (10 foot)
  • Enclosure support feet (4)
  • Power Cord 634061
  • North American power cord set (10 foot)

If the instrument is permanently mounted in an open panel or rack, use electrical metal tubing or conduit.

  • Enclosure Support Kit 634958
  • Enclosure support feet (4)
  • Power Cord/Enclosure Support Kit 654008
Emerson Controller (951C) — S3 diagram (page 20)
Fig. 4

Set switch window for voltage required.

Figure 2-1. Power Supply Board Voltage Select Switches

Emerson Controller (951C) — S3 diagram (page 20)
Fig. 5

S3

Set switch window for voltage required.

Figure 2-2. Temperature Control Board

2-2 Installation Rosemount Analytical Inc. A Division of Emerson Process Management

Page 21 of the Emerson Controller (951C) user manual — INTERIOR EXTERIOR
Page 21 · click to enlarge
Page 21

b. Potentiometric Recorder Connections

Current recorder interconnection cable specs are as follows:

Refer to Figure 2-3 (below), Figure 2-5 (page 2-4) and drawing 654063. Potenti- ometric recorder connections are made on the rear panel. Route the potentiometric re- corder cable through the cable gland in the hole labeled RECORDER OUTPUT and connect to VOLT OUTPUT terminals.

  • Distance the recorder from analyzer: 3000 feet (915 meters).maximum
  • Load resistance: Less than 700 Ohms.
  • Cable (user supplied): Two-conductor, shielded, min. 20 AWG

Potentiometric recorder cable specifications are as follows:

As supplied by the factory, the current output produces a zero of 4 mA. The current output may be adjusted to pro- duce a zero of 0 mA as follows:

  • Distance from recorder to analyzer: 1000 feet (305 meters) maximum
  • Input impedance: Greater than 2000 ohms
  • Cable (user supplied): Two-conductor, shielded, min. 20 AWG
1. Zero the instrument as in Section 34. Adjust R23, the zero-adjust potentiometer on the Power Supply Board, to produce 0 mA current output.
  • Voltage output: 0 to +5 VDC

c. Current Recorder Connections

INTERIOR EXTERIOR

Refer to Figure 2-3 (below), Figure 2-5 (page 2-4) and drawing 654063. Current recorder connections are made on the rear panel. Route the current recorder cable through the cable gland in the hole labeled RECORDER OUTPUT and connect to CUR OUTPUT terminals

Figure 2-3. Cable Gland

Emerson Controller (951C) — INTERIOR EXTERIOR diagram (page 21)
Fig. 6

10 PSI - 17 PSI (70 kPa - 120 kPa)

See Figure 2-5 for connections

Figure 2-4. Rear View of Model 951C (cover removed)

Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-3

Show 4 more labels / short text on this page
Nut Gland Nut · Cable · Case Wall · Fuse
Page 22 of the Emerson Controller (951C) user manual — FUSE
Page 22 · click to enlarge
Page 22

RECORDER

POWER

OUTPUT

J2 Remote Range Change Connections

J2

RANGE 1

RANGE 2

RANGE 3

RANGE 4

RTN

RTN

+24 VDC IN

+24 VDC IN

CHS GND

CHS GND

TB1

L1/HOT

TB1 AC Power Connections

FUSE

L2/NEUT

GND

CUR OUTPUT

VOLT OUTPUT

+ - G + -

TB2

TB2 Recorder Connections TB2 Current Output Connections

Figure 2-5. Rear Panel Wiring Connections

2-4 Installation Rosemount Analytical Inc. A Division of Emerson Process Management

Page 23 of the Emerson Controller (951C) user manual — NOTE
Page 23 · click to enlarge
Page 23

2-5 GAS REQUIREMENTS

2-6 SAMPLE REQUIREMENTS

The instrument requires two gases normally supplied from cylinders. They are:

The sample must be clean and dry before entering the analyzer. In general, before admission to the analyzer, the sample should be filtered to eliminate particles lar- ger than two microns and have a dew point below 90°F (32°C). The factory can provide technical assistance if desired.

a. Air (U.S.P. Breathing Grade)

This is used as both (a) an oxygen source for generation of the ozone required for the chemiluminescence reaction, and (b) a standard gas for zero calibration (nitrogen can also be used). Gas for each purpose must be supplied from a separate cylinder due to different pressure requirements at ozonator and zero inlets.

Proper supply pressure for sample, zero and span gases for the Model 951C is 20 psig (138 kPa).

2-7 GAS CONNECTIONS

b. Span Gas

WARNING

TOXIC AND OXIDIZING GAS HAZARDS

This is a standard gas of accurately known composition, used to set an upscale calibra- tion point. The usual span gas is NO or NO2 in a background of nitrogen.

This instrument generates ozone which is toxic by inhalation and is a strong irritant to throat and lungs. Ozone is also a strong oxidizing agent. Its presence is detected by a characteristic pungent odor.

Emerson Controller (951C) — WARNING diagram (page 23)
Fig. 7

The instrument exhaust contains both ozone and nitrogen dioxide, both toxic by inhalation, and may contain other constituents of the sample gas which may be toxic. Such gases include various oxides of nitrogen, unburned hydrocarbons, carbon monoxide and other products of combustion reactions. Carbon monoxide is highly toxic and can cause headache, nausea, loss of consciousness, and death.

This instrument requires periodic calibration with a known standard gas. See Section 3-3. See also General Precautions for Handling and Storing High Pressure Gas Cylinders, page P-5.

NOTE

For maximum calibration accuracy, the concentration of NO in the span gas should be similar to that in the sample gas. Also, the span gas should be supplied to the rear panel SAMPLE inlet at the same pressure as the sample gas. To ensure constant pressure, a pressure regulator may be utilized immediately upstream from the SAMPLE inlet.

Avoid inhalation of the ozone produced within the analyzer and avoid inhalation of the sample and exhaust products transported within the analyzer. Avoid inhalation of the combined exhaust products at the exhaust fitting.

Each gas used should be supplied from a tank or cylinder equipped with a clean, non-corrosive type, two-stage regulator. In addition, a shut-off valve is desirable. Install the gas cylinders in an area of relatively constant ambient temperature.

Keep all tube fittings tight to avoid leaks. See Section 2-8 for Leak Test Procedure.

Connect rear exhaust outlet to outside vent by a 1/4 inch (6.3 mm) or larger stainless steel or Teflon line. Check vent line and connections for leakage.

Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-5

Page 24 of the Emerson Controller (951C) user manual — 2-8 LEAK TEST
Page 24 · click to enlarge
Page 24
1. Remove plugs and caps from all inlet and outlet fittings. (See Figure 2-4, page 2-3.)

2-8 LEAK TEST

The following test is designed for sample pressure up to 5 psig (35 kPa).

2. Connect EXHAUST outlet to external vent via tubing with O.D. of 1/4-inch (6.3 mm) or larger. Use only stainless steel or Teflon tubing.
1. Supply air or inert gas such as nitrogen at 5 psig (35 kPa) to analyzer sample and air input fittings.
3. Connect external lines from ozonator air and sample sources to corresponding rear panel inlet ports. For sample line, stainless steel tubing is recommended.
2. Seal off analyzer exhaust fitting with a tube cap.
3. Use a suitable test liquid such as SNOOP (P/N 837801) to detect leaks. Cover all fittings, seals, or possible leak sources.
4. Adjust regulator on ozonator air cylinder for output pressure of 20 to 25 psig (138 to 172 kPa). At least 20 psig should be present at rear of analyzer.
4. Check for bubbling or foaming which indicates leakage, and repair as required. Any leakage must be corrected before introduction of sample and/or application of electrical power.
5. Supply sample gas to rear panel SAMPLE inlet at appropriate pressure: 20 psig (138 kPa). The nominal input pressure is 20 psig (138 kPa).

For positive assurance that system is leak free, perform one of the tests in Section 5.0.

2-6 Installation Rosemount Analytical Inc. A Division of Emerson Process Management

Page 25 of the Emerson Controller (951C) user manual — NOTE
Page 25 · click to enlarge
Page 25

SECTION 3 OPERATION

3-1 FRONT PANEL INDICATORS AND CONTROLS

c. Sample Pressure Gauge

The internal SAMPLE pressure (nominally 4 psig, 28 kPa) is adjusted by rotation of the Sample Pressure Regulator. See Figure 3-1 below.

a. Display

The display is a 4-digit liquid crystal de- vice which always displays NOX concen- tration in parts-per-million. Figure 3-1 below.

NOTE

Using the MID ranges (20, 50, 200, and 500 ppm), set the sample pressure regulator to 2.5 psig, 17.5 kPa

b. Range Selection

The Model 951C has twelve customer se- lectable ranges, four LO ranges (10 ppm, 25 ppm, 100 ppm and 250 ppm), four MID ranges (20 ppm, 50 ppm, 200 ppm and 500 ppm), and four HI ranges (100 ppm, 250 ppm 1000 ppm and 2500 ppm). The range is selected by positioning the PPM RANGE Switch (SW1) and the five con- figuration jumpers on the Signal Board to the desired range controlling the recorder output. Refer to Figure 3-2, page 3-2.

d. Ozone Pressure

The OZONE pressure is determined by the pressure regulator of the air supply cylinder. A nominal pressure of 20 to 25 psig (138 to 172 kPa) is recommended. Proper operation is indicated when the front panel OZONE indicator lamp is lit.

NOTE

If ozone lamp does not light, increase pressure slightly by adjusting pressure regulator control on the air cylinder.

The Backlight will blink for values 10% in excess of the range maximum. Moving the switch to the right selects a higher fullscale value and restores the display.

POWER SUPPLY BOARD (See Figure 3-3)

Emerson Controller (951C) — NOTE diagram (page 25)
Fig. 8

CASE HEATER TEMPERATURE CONTROL ASSEMBLY (See Figure 6-4) TEMPERATURE CONTROL BOARD (See Figure 2-2)

Ozone Indicator Lamp (Signal Board DS2) Range1 Control (Signal Board R101)

Figure 3-1. Model 951C Controls, Indicators and Adjustments

Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-1

Page 26 of the Emerson Controller (951C) user manual — R25 R43
Page 26 · click to enlarge
Page 26

e. Zero and Span Potentiometers

f. Ozone Interlock

See Figure 3-1 (page 3-1) and Figure 3-2 below. Screwdriver access holes through the front panel allow adjustments of the ZERO, RANGE1 and RANGE2 potenti- ometers (R100, R101 and R102 on the Signal Board).

The ozone-producing UV lamp will not ig- nite or stay lit unless adequate air pres- sure is present at the AIR inlet (see Figure 2-4, page 2-3). Nominal setpoint pressure is 20 to 25 psig.

PPM Range Select Switch (Figure 3-6, page 3-3) Lo/Mid/Hi Selection Jumpers (Figure 3-6, page 3-3)

R25 R43

Emerson Controller (951C) — R25 R43 diagram (page 26)
Fig. 9

Figure 3-2. Signal Board

Emerson Controller (951C) — R25 R43 diagram (page 26)
Fig. 10

Figure 3-3. Signal Board Test Points

3-2 Operation Rosemount Analytical Inc. A Division of Emerson Process Management

Show 5 more labels / short text on this page
\ · Electrometer Assembly · Zero (R100) · Range 1 (R101) · Range 2 (R102)
Page 27 of the Emerson Controller (951C) user manual — 3-2 STARTUP PROCEDURE
Page 27 · click to enlarge
Page 27

3-2 STARTUP PROCEDURE

c. Set Range2 Selection jumpers to

The following are detailed instructions on startup and calibration.

Range 4, 2500ppm as shown in Figure 3-8 below.

1. Supply electrical power to the analyzer. The analyzer will require approximately two hours for temperature equilibration before calibration.
Emerson Controller (951C) — 3-2 STARTUP PROCEDURE diagram (page 27)
Fig. 11
2. On Signal Board, Figure 3-2, page 3-2:

JP1 10 20 100 JP2 25 50 250 JP3 100 200 1000 JP4 250 500 2500

a. Using a small flat screwdriver, set

PPM RANGE Switch (Figure 3-6 below) to the range that will be used for sample analysis.

Figure 3-8. Range 2 Selection Jumpers

Emerson Controller (951C) — 3-2 STARTUP PROCEDURE diagram (page 27)
Fig. 12

Hi Range 1 10 20 100 2 25 50 250 3 100 200 1000 4 250 500 2500 Other Remote Remote Remote

Figure 3-6. PPM Range Select Switch

b. Set Configuration Jumper Settings to

desired position as shown in Figure 3-7 below.

Emerson Controller (951C) — 3-2 STARTUP PROCEDURE diagram (page 27)
Fig. 13

Figure 3-7. Configuration Jumper Settings

Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-3

Page 28 of the Emerson Controller (951C) user manual — NOTE
Page 28 · click to enlarge
Page 28

b. Adjust SAMPLE Backpressure Regula-

3. Establish correct pressure for air by the following:

tor so internal SAMPLE Pressure Gauge indicates the value appropriate to the desired operating range.

a. Adjust OZONE Pressure Regulator so

that OZONE Pressure Gauge indi- cates 20 to 25 psig (138 to 172 kPa).

NOTE

Inability to obtain a flow of one liter per minute at the EXHAUST outlet usually indicates insufficient sample supply pressure at the SAMPLE inlet. Use a 2400 cc flowmeter (i.e., Brooks P/N 1350) at the EXHAUST outlet to measure flow.

b. To establish correct pressure of zero gas,

supply zero gas to rear panel SAMPLE inlet. Note reading on internal SAMPLE Pressure Gauge. It should be the same as the nominal 4 psig (28 kPa) for the Lo and Hi Ranges and 2.5 psig for the Mid Range SAMPLE pressure indicated on the internal SAMPLE pressure gauge. This should remain constant when the analyzer input SAMPLE is switched from calibration gas standard to a zero gas standard. This may be assured by setting the delivery from the SAMPLE and the zero gas cylinder of span gas cylinder to the same value of delivery pressure, nominally 20 psig (138 kPa). If not, ad- just output pressure regulator on zero gas cylinder as required.

5. Establish correct flow of upscale standard gas by the following:

a. Supply upscale standard gas to rear

panel SAMPLE inlet.

b. Note reading on internal SAMPLE

Pressure Gauge. It should be the same as in Step 3b.

NOTE

Supply pressures for sample and upscale standard gases must be the same. Otherwise, readout will be in error.

4. Establish correct pressure of sample gas by the following:

a. Supply sample gas to rear panel

SAMPLE inlet.

The analyzer is now ready for calibration.

3-4 Operation Rosemount Analytical Inc. A Division of Emerson Process Management

Page 29 of the Emerson Controller (951C) user manual — 3-2 STARTUP PROCEDURE
Page 29 · click to enlarge
Page 29
Emerson Controller (951C) — 3-2 STARTUP PROCEDURE diagram (page 29)
Fig. 14
Emerson Controller (951C) — 3-2 STARTUP PROCEDURE diagram (page 29)
Fig. 15
Emerson Controller (951C) — 3-2 STARTUP PROCEDURE diagram (page 29)
Fig. 16
Emerson Controller (951C) — 3-2 STARTUP PROCEDURE diagram (page 29)
Fig. 17

Figure 3-9. Power Supply Board

Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-5

Page 30 of the Emerson Controller (951C) user manual — 3-5 CONVERTER TEMPERATURE ADJUSTMENT PROCEDURE
Page 30 · click to enlarge
Page 30

3-3 CALIBRATION

5. If necessary, increase sensitivity by raising photomultiplier voltage. Adjust R30 on the Power Supply Board (Figure 3-9) clockwise to increase (negative) the photomultiplier high voltage and sensitivity. This will interact with zero. Repeat Zero Calibration and Upscale Calibration (through step 3).

a. Zero Calibration

1. On the Signal Board (Figure 3-2, page 3-2), set PPM RANGE Switch (Figure 3-6, page 3-3) for the same range that will be used during sample analysis. Set front panel RANGE1 Control at about mid-range.
6. If necessary increase the upscale readings on the LCD display by adjusting potentiometer R43 until the display shows the correct span gas readings. Now, adjust R 25 so that reading on display and the recorder is equal to the known parts-per-million concentration of NO X in the span gas.
2. Supply zero gas to rear panel SAMPLE inlet.
3. Using the MID Range set the sample pressure regulator to 2.5 psig. Using the Lo or Hi ranges, set the sample pressure to 4.0 psig
4. After a stable reading is reached, adjust the zero by inserting a screwdriver in the ZERO slot on the front panel of the analyzer and turning until zero reading is obtained.

3-4 ROUTINE OPERATION

After calibrating analyzer per Section 3-3, supply sample to SAMPLE inlet. Set PPM RANGE Switch (Figure 3-6, page 3-3) in ap- propriate position. The instrument will now continuously analyze the sample stream.

b. Upscale Calibration

The Model 951C is designed for continuous operation. Normally, it is never turned off ex- cept for servicing or for a prolonged shut- down.

1. On the Signal Board (Figure 3-2, page 3-2) set Upscale Calibration PPM RANGE Switch (Figure 3-6, page 3-3) to the position appropriate to the particular span gas.

NOTE

2. Supply upscale standard gas of accurately known NO X content to rear panel SAMPLE inlet.

During periods of shutdown, turn off the ozone lamp by shutting off the input air source.

3. Using the MID Range set the sample pressure regulator to 2.5 psig. Using the Lo or Hi ranges, set the sample pressure to 4.0 psig

3-5 CONVERTER TEMPERATURE ADJUSTMENT PROCEDURE

Once the appropriate high voltage and elec- tronic gain have been selected such that the named calibration gas value is indicated by the Model 951C, the instrument is ready for adjustment of the converter temperature.

4. Adjust front panel RANGE1 Control so that reading on display or recorder is equal to the known parts-per-million concentration of NO X in the span gas. If the correct reading is not initially attainable by adjustment of the front panel RANGE1 Control, make the electronic adjustment in Step 5.

The vitreous carbon converter used in this analyzer has a low surface area which gradu- ally increases during high temperature opera- tion of the converter material.

3-6 Operation Rosemount Analytical Inc. A Division of Emerson Process Management

Page 31 of the Emerson Controller (951C) user manual — TEMPERATURE
Page 31 · click to enlarge
Page 31

Initially, the temperature of the peak of the converter efficiency starts at a relatively high value because significant heat must be sup- plied to make the converter active enough to reduce the input nitrogen dioxide to nitric ox- ide at the required 95% level. During the op- eration of the analyzer, the temperature of the peak will fall as the surface area of the con- verter is increased and less external energy is required to cause adequate conversion.

3. Refer to Power Supply Board Figure 3-9, page 3-5. Turn the converter temperature adjust potentiometer (R9 CONV HTR) one turn counterclockwise from the setting established at the factory, and allow fifteen minutes for operation at the new lower temperature setpoint. Recheck the response and note the value for later use.

4. Increase the temperature of the converter by rotating the converter temperature ad- just potentiometer (R9 CONV HTR) one quarter turn clockwise, wait fifteen min- utes for thermal equilibrium and then re- measure the NO2 calibration gas value. Note its value. Repeat this procedure of one quarter turn adjustments of the poten- tiometer, waiting for thermal stability and determination of the calibration gas value until either a 95% value is obtained or the final one quarter turn adjustment gives an efficiency increase of less than one per- cent.

In extreme cases, where converter re-profiling has not been conducted, the converter is so active that it not only reduces nitrogen dioxide to nitric oxide, but it reduces the nitric oxide produced to nitrogen, which is not detected by the chemiluminescence reaction. The remedy in this case is to adjust the converter tempera- ture to a lower value to improve the converter efficiency.

It is important that the converter temperature be periodically profiled to assure that it is run- ning at its peak efficiency. An interval of one week is recommended. The nominal range of operational temperatures for the converter is 275°C to 400°C (527°F to 750°F). The operat- ing temperature of the converter may be con- veniently checked on the Power Supply Board (Figure 3-9, page 3-5) by momentarily de- pressing switch CONV TEMP CHECK (S4) while monitoring the resistance across termi- nals TP1 and TP2.

5. Decrease the temperature of converter operation by rotating the converter temperature adjust potentiometer (R9 CONV HTR) one eighth of a turn counterclockwise. This places the converter at a temperature suitable for low ammonia interference and efficient NO 2 conversion. Re-measure the indicated converter temperature and compare it to the initially recorded value.

Table 3-1 allows for conversion of the ob- served resistance to the operating tempera- ture for the converter.

TEMPERATURE

RESISTANCE

( ° C)

(Ohms) 0 400 25 438 100 552 200 704 250 780 300 856 350 932 400 1008 450 1084

Follow this procedure to optimize the operat- ing temperature of the converter:

1. Power instrument and allow it to stabilize at operating temperature (one to two hours). Measure the operating temperature of the converter by the technique described above. Note the value for future reference.
2. Admit a calibration gas of known (NO 2 ) concentration into the analyzer and note the concentration value determined when the full response has been achieved.

Table 3-1. Resistance of Converter Temperature

Sensor vs. Temperature

Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-7

Page 32 of the Emerson Controller (951C) user manual — NOTE
Page 32 · click to enlarge
Page 32

NOTE

Converter temperature is not a direct measure of converter efficiency. Temperature measurement is for reference purposes only.

3-8 Operation Rosemount Analytical Inc. A Division of Emerson Process Management

Page 33 of the Emerson Controller (951C) user manual — 3-7 RECOMMENDED CALIBRATION FREQUENCY
Page 33 · click to enlarge
Page 33

3-6 MEASUREMENT OF CONVERTER EFFICIENCY

consume nitrogen dioxide as it delivers sample to the analyzer. Losses of nitro- gen dioxide in the pump will be reported as converter inefficiency.

It is the responsibility of the user to measure efficiency of the NO2-to-NO converter during initial startup, and thereafter at intervals ap- propriate to the application (normally once a month).

4. Operate the analyzer and continue to sample the diluted nitric oxide sample for a period of at least thirty minutes. If the nitrogen dioxide to nitric oxide conversion is at the 100% level, the instrument response will be stable at the highest value noted.

Section 3-5 optimizes the operating tempera- ture of the converter. It also serves as an effi- ciency check if the concentration of NO2 in the calibration gas is documented accurate rela- tive to National Institute of Standards and Technology (NIST) Reference Materials. If the concentration of the nitrogen dioxide calibra- tion gas is not known accurately, this proce- dure still serves to adequately provide the correct converter operating temperature.

5. If the response at the end of the thirty mi- nute period decreases more than 2.0 per- cent of the highest peak value observed, the system is not acceptable and correc- tions must be made before repeating the check. If it is determined that observed subnormal conversion efficiencies are real, and not due to errors introduced by nitrogen dioxide consumption in the sam- ple pump or other parts of the sample handling system, verify that the converter is peaked at the optimum temperature be- fore replacing with a new converter.

If the only available known standard is the ni- tric oxide calibration standard, the following procedure may be performed. This procedure checks converter efficiency through the utiliza- tion of gas-phase oxidation of nitric oxide into nitrogen dioxide over a range of nitrogen diox- ide concentrations. This technique is ab- stracted and adapted from 40 CFR, Pt. 60, App. A, Method 20, Paragraph 5.6.

3-7 RECOMMENDED CALIBRATION FREQUENCY

1. Select the appropriate instrument range.

After initial startup or startup following a shut- down, the analyzer requires about two hours for stabilization before it is ready for calibra- tion. Maximum permissible interval between calibrations depends on the analytical accu- racy required, and therefore cannot be speci- fied. It is recommended that initially the instrument be calibrated at least once every 8 hours. This practice should continue until ex- perience indicates that some other interval is more appropriate.

2. Admit a nitric oxide in nitrogen NIST traceable calibration gas of a value between 45% and 55% of the instrument range selected to a clean, evacuated, leak tight Tedlar bag. Dilute this gas approximately 1:1 with a 20.9% oxygen, purified air.
3. Immediately attach the bag outlet to the input of the pump supplying pressurized gas to the analyzer. It is important to use a sample delivery pump which does not

Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-9

Page 34 of the Emerson Controller (951C) user manual — 3-6 MEASUREMENT OF CONVERTER EFFICIENCY
Page 34 · click to enlarge
Page 34
Page 34 of the Emerson Controller (951C) user manual — 3-6 MEASUREMENT OF CONVERTER EFFICIENCY
Page 35 of the Emerson Controller (951C) user manual — UV 3O 2 → 2O 3
Page 35 · click to enlarge
Page 35

SECTION 4

THEORY

4-1 NITRIC OXIDE DETERMINATION BY CHEMILUMINESCENCE METHOD

The flow rate of the selected gas into the reaction chamber is controlled by a back pressure regulator inside the ana- lyzer. It provides an adjustable, con- trolled pressure on the upstream side, where gas is supplied to the calibrated, flow-limiting sample capillary. The regu- lator is adjusted for appropriate reading on the internal SAMPLE Pressure Gauge. For operation at NO and NO2 levels below 250 ppm, correct setting on the SAMPLE Pressure Gauge is 4 psig (28 kPa). This results in a flow of approximately 60 to 80 cc/min to the re- action chamber.

The chemiluminescence method for detection of nitric oxide (NO) is based on its reaction with ozone (O3) to produce nitrogen dioxide (NO2) and oxygen (O2). Some of the NO2 molecules thus produced are initially in an electronically excited state (NO2*). These revert immediately to the ground state, with emission of photons (essentially red light).

The reactions involved are:

NO + O3 → NO2* + O2

NO2* → NO2 + Red Light

Excess sample is discharged with the effluent from the reaction chamber via the EXHAUST outlet. Bypass flow is set by the restrictor at 1 L/min (nominal) to ensure proper functioning of the SAM- PLE Pressure Regulator and rapid sys- tem response. Excessive changes, on the order of 5 psig (35 kPa), in the pressure of the sample or standard gas will affect the bypass flow rate and can affect accuracy.

As NO and O3 mix in the reaction chamber, the intensity of the emitted red light is proportional to the concentration of NO.

(Any NO2 initially present in the sample is re- duced to NO by a heated bed of vitreous carbon through which the sample is passed before be- ing routed to the reaction chamber.)

The intensity of the emitted red light is meas- ured by a photomultiplier tube (PMT), which produces a current of approximately 3 X 10-9 amperes per part-per-million of NO in the reac- tion chamber.

b. Ozone Generation

Suitably pressurized air from an exter- nal cylinder is supplied to the rear panel AIR inlet. The proper pressure setting is 20 to 25 psig (138 to 172 kPa). Within the ozone generator, a portion of the oxygen in the air is converted to ozone by exposure to an ultraviolet lamp. The reaction is:

4-2 ANALYZER FLOW SYSTEM

The analyzer flow system is shown in drawing 2-0 and 3-0. Its basic function is to deliver regu- lated flows of sample, calibration gas, or zero gas and ozonized air to the reaction chamber. The discharge from the reaction chamber flows from the analyzer via the EXHAUST outlet.

UV 3O 2 → 2O 3

From the generator, the ozonized air flows into the reaction chamber for use in the chemiluminescence reaction.

a. Flow of Sample, Standard Gas or Zero

Gas to Reaction Chamber

Suitably pressurized sample, standard gas or zero gas is supplied to the rear panel SAMPLE inlet.

Rosemount Analytical Inc. A Division of Emerson Process Management Theory 4-1

Page 36 of the Emerson Controller (951C) user manual — 4-3 SIGNAL CONDITIONING AND DISPLAY
Page 36 · click to enlarge
Page 36

4-3 SIGNAL CONDITIONING AND DISPLAY

cluded jumpers, this is used for higher range selection (ie) 100 to 2500 ppm NOx.

A block diagram of the signal-processing elec- tronics is shown in Figure 4-1, page 4-3.

Gain Amplifier U14 The precision amplifier U14 allows 2 se- lectable gains, on for 10, 100, and 1000 spans and the other for 25, 250, and 2500 spans.

The signal conditioning and display board pro- vides the following functions:

  • Signal conditioning circuit consisting of U14

Intermediate gain ranges as required by the 951C Lo Range: 10, 25, 100, 250 ppm NOx, Mid Range: 20, 50, 200, 500 ppm NOx or Hi Range 100, 250, 1000, 2500 ppm NOx are obtained by interpo- lating between the requisite ranges us- ing Boolean Logic and Analog Switches.

  • An analog to digital converter U15.
  • Display device U15
  • Range Control circuits U9 and U12
  • Post signal amplifier and output amplifier circuit U5

Range 1 and Range 2 selection The signal output of U14 is adjusted at the front panel for instrument calibration purposes. Any second range may be chosen by placing a jumper on JP1, JP2, JP3, or JP4. When the pro- grammed range is selected, the K2 re- lay toggles and allowing that range to be calibrated. Note! If no jumper is pre- sent only range 1 may be calibrated

  • Display/Backlight blink control U5 and U6
  • Range conditions.
  • Remote Control Circuits

All of the above functions are on a single board located at the front of the instrument. Certain control requirements (ie) range calibration and zero offset adjustment are available as screw- driver adjustments via the front panel. A digital liquid crystal display is mounted on the rear side of the board for data display purposes as are control potentiometers.

Post Amplifiers U5 The conditioned signal from U14 and the range calibration potentiometers R101 and R102 are provided to U5. This amplifier has switch controlled feed back resistors that permit gain selec- tions of any range, as determined by the range switch and associated logic. An Adjustment potentiometer R25 al- lows a small correction for inter range calibration purposes.

a. Circuit Functions

Signal conditioning amplifiers Boards assembly 6A00326G01/02 is a high input impedance electrometer amplifier board.

The output of U14 is provided to the Analog to Digital converter for digitiza- tion and display purposes. Potentiome- ter R43 allows adjustment of the signal to the ADC allowing concordance of dis- play and recorder signals if required. Normally the ADC should be correctly calibrated during the factory test proce- dure and should not require further ad- justment. Potentiometer R25 allows

Current output from the detector unit is con- verted to voltage by the electrometer ampli- fier and then further amplified by post amplifier.

The electrometer amplifier gain may be re- duced by a factor of 10 by shorting JP1 and by shorting pins 1 and 2 on P1 with the in-

4-2 Theory Rosemount Analytical Inc. A Division of Emerson Process Management

Page 37 of the Emerson Controller (951C) user manual — 4-4 ANALYZER THERMAL SYSTEM
Page 37 · click to enlarge
Page 37

recorder output adjustment to match dis- play.

Connecting any input terminal 1 thru 4 to ground terminals 9 or10 will select one of the four ranges.

U5 is the recorder output amplifier. The out- put signal may be trimmed by R25 to pre- cisely match the recorder calibration. Note that this adjustment does not affect or change the instrument display,

For remote operation a 24V power sup- ply should be connected to terminals 5,6 and 7,8, negative or low side is connected to 5, 6. Positive or high side to 7,8. For protection purposes the high side is fused. Note! The local range switch SW1 must be in the fifth posi- tion (position Five) to disconnect lo- cal control and provide the return for remote control.

Analog to Digital Converter U15 is a special integrated circuit with a built in ADC and LCD Drivers. The signal data is digitized and provided in the correct format to the liquid crystal display (LCD).

For local operation Set the range switch SW1 between setting 1 thru 4. An ex- ternal 24V supply is not required. SW1 is located on the top of the signal board.

Display/Backlight and over range blink Circuits The display circuits comprise of U15, no ad- justments are required for these functions

4-4 ANALYZER THERMAL SYSTEM

The blink or over range display function uses U5, U6, and U9.

The Analyzer Thermal System is shown in Figure 4-2, page 4-3. Its basic function is to provide a stable thermal environment for the PMT.

U5 is an analog comparator that senses an over range condition and turns on U6. U6 sets the blink rate for the Backlight.

The temperature of the PMT must be held within a half-degree band at approximately 18°C if it is to produce a useful signal for low concentrations of NOX. This is accom- plished by means of a solid-state cooler which houses the PMT. The heat which is radiated from the cooler is carried away by the Cooler Fan.

Remote Control Circuits The Model 951C is provided with a fully iso- lated remote control interface. Optical isola- tors and a remote 24V power supply ensure that no direct return path exists between the users system and the 951C When in the remote mode of operation.

Relay K1 switches between the internal range switch control (SW1) and the External Remote Range Control. When SW1 is on setting 5 it causes K1 to toggle and a fully isolated system is achieved.

The solid-state cooler must work against a relatively constant load in order to maintain the temperature of the PMT. This load is produced by a case heater and exhaust fan which control the temperature inside the case within a one-degree band (approxi- mately 50°C for ambient temperatures from 4°C to 40°C).

Optical Isolators U7, U8, U10, and U13 con- nects via a ribbon cable to the remote con- nector panel at the rear of the 951C. A ten terminal barrier strip provides connection for remote range selection.

The electronics which support the Analyzer Thermal System and the NO2-to-NO Con- verter are contained on the Power Supply Board.

Rosemount Analytical Inc. A Division of Emerson Process Management Theory 4-3

Page 38 of the Emerson Controller (951C) user manual — SECTION 4
Page 38 · click to enlarge
Page 38

PMT Electrometer Gain Amplifier Span Amplifier

Span Adjustments R101, R102

Figure 4-1. Analyzer Block Diagram

Rosemount Analytical Inc. A Division of Emerson Process Management Theory 4-5

Show 15 more labels / short text on this page
Signal/Control Board · Display · Photomultiplier · Tube · Potentiometric Output · Voltage Output · Zero Control · Range Switch · Control · High Voltage · Power Supply Board · Isolated Current Output · Supply · Voltage-to-Current · Converter
Page 39 of the Emerson Controller (951C) user manual — EXHAUST FAN
Page 39 · click to enlarge
Page 39

INLET VENT HOLES

Top View of Analyzer

EXHAUST FAN

SOLID-STATE COOLER

PMT

CASE HEATER

FRONT PANEL

Figure 4-2. Analyzer Thermal System

4-6 Theory Rosemount Analytical Inc. A Division of Emerson Process Management

Cooling Fins · Cooler Fan · Fan Heater

Page 40 of the Emerson Controller (951C) user manual — SECTION 4
Page 40 · click to enlarge
Page 40

Rosemount Analytical Inc. A Division of Emerson Process Management Theory 4-7

Page 41 of the Emerson Controller (951C) user manual — NOTE
Page 41 · click to enlarge
Page 41

SECTION 5 ROUTINE SERVICING

a. Display Fullscale Span Adjustment

WARNING ELECTRICAL SHOCK HAZARD

If a recorder is used, and has been prop- erly zeroed, it should agree with the dis- play reading. If not, obtain agreement by adjustment of R25 on the Signal Board (see Figure 3-1 (page 3-1) and Figure 3-2 (page 3-2). If agreement cannot be reached, check the recorder. If the re- corder is functioning properly, replace the Signal Conditioning Board.

Servicing requires access to live parts which can cause death or serious injury. Refer servicing to qualified personnel.

WARNING

INTERNAL ULTRAVIOLET LIGHT HAZARD

Ultraviolet light from the ozone generator can cause permanent eye damage. Do not look directly at the ultraviolet source in ozone generator. Use of ultraviolet filtering glasses is recommended.

b. Overall Sensitivity

Principal factors that determine overall sensitivity of the analyzer are the follow- ing: (a) sample flow rate to the reaction chamber, (b) sensitivity of the photomulti- plier tube (PMT), and (c) PMT high volt- age. If specified fullscale readings are unobtainable by adjustment of the SPAN Control, sensitivity is subnormal. The cause of reduced sensitivity may be in ei- ther the flow system (See Section 5-2, page 5-3) or the electronic circuitry (See Section 5-6, page 5-10).

NOTE

The photomultiplier tube must not be exposed to ambient light. If the photomultiplier tube is ex- posed to light while the power is on, either through a loose fitting on the reaction chamber or any other leak, it will be destroyed. If exposed to ambient light with the power off, the tube will be noisy for some period of time. Unless appro- priate precautions are observed, light can strike the tube upon removal of fittings from the reac- tion chamber. 5-1 SYSTEM CHECKS AND ADJUSTMENTS

If either the High Voltage Board or the Phototube/Reaction Chamber Assembly has been replaced, a readjustment of the high voltage will probably be required to obtain the correct overall sensitivity. Ad- just R30 on the Power Supply Board (see Figure 3-1 page 3-1 and Figure 3-9 page 3-5) clockwise to increase (negative) the photomultiplier high voltage and sensitiv- ity, or counterclockwise to decrease (negative) the voltage and sensitivity. The adjustment range is about -650 V to -2100 V for the regulated DC voltage ap- plied to the photomultiplier tube. Nominal setting is -1100 volts. However, the volt- age should be adjusted as required for overall system sensitivity.

The following procedures may be used to de- termine the cause of unsatisfactory instrument performance, or to make adjustments follow- ing replacement of components. If a recorder is available, use it for convenience and maxi- mum accuracy in the various tests.

Rosemount Analytical Inc. A Division of Emerson Process Management Troubleshooting 5-1

Page 42 of the Emerson Controller (951C) user manual — 5-2 SERVICING FLOW SYSTEM
Page 42 · click to enlarge
Page 42

c. Ozone Output

display reading with adjustment of the ZERO Control. If this cannot be accom- plished, the cause must be found and cor- rected. The fault may be in either the electronic circuitry or the sample flow sys- tem.

WARNING

TOXIC GAS HAZARD

Use extreme caution in troubleshooting the ozone generator. Ozone is toxic.

First, establish proper performance of the electronic circuitry. Turn on analyzer power. Verify that ZERO Control and am- plifier are functioning properly. Then, check for excessive photomultiplier dark current and/or contamination of the reac- tion chamber or sample flow system as follows:

To check for adequate output from the ozone lamp, a convenient technique is to calibrate the analyzer on a high level NO standard such as 250 ppm NO at the nominal 4.0 psi internal sample pressure setpoint, and note the reading. The sam- ple pressure setpoint is then sequentially set to pressures of 3.0, 2.0, and 1.0 psi after a stable span gas reading has ob- tained at the higher pressure setpoint. The span gas value will change as the pressure is changed. The difference in span gas value between two successive sample pressure levels should be ap- proximately the same for the 4.0 to 3.0, 3.0 to 2.0, and 2.0 to 1.0 pressure steps.

  • Excessive Photomultiplier Dark Current

To check, shut off all flow to the ozone generator. Turn off ozone generator. Supply cylinder air to rear panel SAMPLE inlet. Note response on display or recorder. If back- ground is still excessive, possible causes are:

If the size of the span gas value differ- ence increases as the sample pressure is lowered, the analyzer output is limited by the amount of ozone production from the lamp and the two additional checks should be made. First, verify that the sample flow (not including bypass) does not exceed the nominal 60 to 80 cc/min, at 4.0 psi internal sample pressure. Sec- ond, substitute another lamp to see if the ozone output is increased.

leakage of ambient light to photomul- tiplier tube

electrical leakage in socket assembly

  • Contamination of Reaction Chamber or Sample Flow System.

See Section 5-4a, page 5-7.

If no other ozone lamp is available, the analyzer sample input pressure may be reduced to the pressure where the ozone limitation is not present. If the lamp output is low and the sample pressure is reduced to restore operation to the condition where ozone limitation is not occurring, some degradation in analyzer response time characteristics may occur.

5-2 SERVICING FLOW SYSTEM

To facilitate servicing and testing, the Model 951C has front drawer access.

Drawing 654090 shows flow system details, including fittings, thread specifications and connecting tubing.

d. Background Current

a. Cleaning Sample Capillary

With zero air supplied to rear panel SAM- PLE inlet, excessive background current is evidenced by the inability to obtain zero

If clogging of sample capillary is sus- pected, measure flow rate as described below.

5-2 Routine Servicing Rosemount Analytical Inc. A Division of Emerson Process Management

defective photomultiplier tube

Page 43 of the Emerson Controller (951C) user manual — 5-3 PHOTOMULTIPLIER TUBE/REACTION CHAMBER
Page 43 · click to enlarge
Page 43
1. Turn off instrument power and shut off all gases.

NOTE

Do not over-tighten capillary internal fitting, as over-tightened fittings may restrict the sample flow.

2. Refer to Figure 6-1 (page 6-4) and Figure 6-2 (page 6-5). Cover and shade the fittings on the reaction chamber with a dark cloth or other light-shielding material. Remove the fitting associated with the sample capillary and place a cap over the open fitting to prevent entry of stray light.

b. Ozone Restrictor Fitting

With instrument power off, supply suitable test gas (dry nitrogen or air) to rear panel AIR inlet. Cover photomultiplier housing with a dark cloth. At the fittings on the re- action chamber, disconnect the ozone tube and place a cap over the open fitting to prevent entry of ambient light. Connect a flowmeter to open end of ozone tube. Adjust the OZONE Pressure Regulator so that the OZONE Pressure Gauge indi- cates normal operating pressure of 20 to 25 psig (138 to 172 kPa). Verify that test flowmeter indicates an appropriate flow of 500 to 600 cc/min for 20 psig.

NOTE

If the opened fitting is inadvertently exposed to ambient light, the instrument will temporarily give a highly noisy background reading. If so, this condition may be corrected by leaving the instrument on, with high voltage on, for several hours. If high voltage is on during exposure, the photomultiplier tube will be destroyed.

Subnormal flow indicates clogging in the flow path that supplies air to the ozone generator. This path contains a Restrictor (P/N 655519), consisting of a metal fitting with internal fritted (metal membrane) re- strictor to reduce pressure. The fitting is upstream from the inlet port of the ozone generator. If the internal restrictor be- comes plugged, the assembly (P/N 655519) must be replaced as it cannot normally be cleaned satisfactorily.

3. With instrument power off, supply suitable test gas (dry nitrogen or air) to rear-panel SAMPLE inlet.
4. Connect a flowmeter to open end of sample capillary. Adjust internal SAMPLE Pressure Regulator to normal operating setting of 4 psig (28 kPa). Verify that flowmeter indicates appropriate flow of 60 to 80 cc/min.
5. If flow is correct, restore analyzer to normal operation.

5-3 PHOTOMULTIPLIER TUBE/REACTION CHAMBER

6. If flow is low, the capillary requires cleaning or replacement (Proceed with the step 7 below).

This assembly consists of the photomultiplier tube and socket, the thermoelectric cooler, and the reaction chamber. Refer to Figure 6-1 (page 6-4) for location and details of mount- ing. Refer to Figure 6-2 (page 6-5) for infor- mation on the assembly.

7. Clean capillary with denatured alcohol, and purge with dry nitrogen or air for one minute. Reconnect capillary.

The assembly must be removed from the ana- lyzer in order to clean the reaction chamber or to replace the photomultiplier tube.

8. With the photomultiplier still covered, slowly insert the free end of the capillary into the corresponding fitting on the reaction chamber. Push the capillary in until it touches bottom against the internal fitting. Then tighten fitting 1/4 turn past finger tight.

Rosemount Analytical Inc. A Division of Emerson Process Management Routine Servicing 5-3

Page 44 of the Emerson Controller (951C) user manual — NOTE:
Page 44 · click to enlarge
Page 44

a. Removal

If sample gas is properly filtered, the reac- tion chamber should not require frequent cleaning. In event of carryover or con- tamination, however, the chamber should be disassembled to permit cleaning the quartz window and the optical filter. The following procedure is recommended.

To remove the photomultiplier tube/reaction chamber assembly from the analyzer, do the follow:

1. Disconnect power from the analyzer.
2. Release pressure from SAMPLE and AIR supplies.
1. Cover and shade the Reaction Chamber/Photomultiplier Assembly with a dark cloth or other light-shielding material.
3. Unplug the electrical cable from the Power Supply PC Board.

NOTE

4. Disconnect the high-voltage cable and the signal cable from the left side of the assembly. Note the two mounting screws just below the connectors.

Always wear surgical rubber gloves when handling the reaction chamber to prevent contamination from handling.

2. Note the orientation of the fittings. Slowly rotate and withdraw the reaction chamber from the housing. Ensure that no light strikes the photomultiplier tube.
5. Uncouple the sample and ozone capillaries and the exhaust line from the right side of the assembly. Note the two mounting screws just below the fittings.
3. Unscrew plastic end cap, thus freeing the quartz window and the red plastic optical filter. Note the sequence in which these are assembled.
6. Loosen the screws described in steps 4 and 5 above.
7. Lift the assembly from the analyzer.
4. Clean the reaction chamber by the appropriate one of the following two methods, standard or alternate. The standard method is applicable in most cases. The alternate method is applicable when the instrument has shown high residual fluorescence. That condition is indicated by high residual currents on a zero gas and high differentials between zero gas readings obtained with the ozone lamp on and off.
8. Replace the assembly by reversing the order of steps 1 through 7 above.

b. Cleaning Reaction Chamber

NOTE:

Photomultiplier tube will be permanently damaged if exposed to ambient light while powered with high voltage. Photomultiplier tube will develop temporary electronic noise if exposed to ambient light with high voltage OFF. A temporary noisy condition may be corrected by leaving instrument on, with high voltage on, for several hours. The required recovery time depends on intensity and duration of the previous exposure. Noise level on the most sensitive range usually drops to normal within 24 hours.

Using a stiff plastic bristle brush, such as a toothbrush, scrub the Teflon surface and gas ports of the reaction chamber with clean distilled water and Alconox* de- tergent (P/N 634929). Alconox detergent is included in the shipping kit provided with the Model 951C NOX Analyzer, and is available from Sargent-Welch Scientific

5-4 Routine Servicing Rosemount Analytical Inc. A Division of Emerson Process Management

Standard Cleaning Procedure

Page 45 of the Emerson Controller (951C) user manual — WARNING
Page 45 · click to enlarge
Page 45

Company under its catalog number S-195650-A.

c. Photomultiplier Tube and Housing

The photomultiplier tube operates at high DC voltages (nominal setting is -1100 volts) and generates small currents that are highly amplified by the sig- nal-conditioning circuitry. It is therefore important that ambient humidity and con- densed water vapor be excluded from the interior of the photomultiplier housing. Ambient humidity can result in electrical leakage, observed as abnormally high dark current. Water vapor or condensed moisture in contact with the photomulti- plier tube may result in an abnormally high noise level during instrument readout on zero air or upscale standard gas.

Using Alconox and clean, soft facial tissue (NOT industrial wipes), carefully clean the quartz window. Vigorously flush reaction chamber and quartz window with clean distilled water. Blow out all possible water from internal passages of reaction cham- ber. Dry reaction chamber and quartz window in a warm oven at 125°F to 150°F (52°C to 66°C) for 30 to 45 minutes or purge-dry the parts with dry cylinder air or nitrogen to eliminate all moisture.

WARNING

ACID HAZARD

The Photomultiplier Tube/Reaction Chamber Assembly incorporates several features for exclusion of humidity and moisture. The photomultiplier socket as- sembly is potted with high impedance sili- cone rubber compound and is sealed from external influences with epoxy and rubber gasket material. The socket as- sembly and the reaction chamber are sealed with O-rings into opposite ends of the tubular photomultiplier housing. The socket end of the housing may be sealed with either one or two O-rings, depending on the length of the phototube.

Hydrochloric acid (HC1) is a strong acid. It is irritating to the skin, mucous membranes, eyes and respiratory tract. Direct contact causes severe chemical burns.

Avoid Contact with eyes and skin and avoid breathing fumes. Use in hood or well ventilated place. Wear goggles, rubber gloves and protective clothing.

Alternate Cleaning Procedure - For High Residual Fluorescence

Holding the reaction chamber by the tube fittings, and using appropriate caution, immerse the white Teflon part of the chamber in 50% concentrated Reagent Grade hydrochloric acid. After five min- utes, rinse thoroughly with de-ionized wa- ter, then air dry as in the standard cleaning method above.

d. Replacement of Photomultiplier Tube

The photomultiplier tube assembly must be removed from the housing in order to replace the tube. To remove, do the fol- lowing:

1. Note the orientation of the connectors.

Place parts in position and press on end-cap so that mating threads engage properly, without cross threading. Turn mating parts in one continuous motion un- til the parts mesh. Do not over-torque.

2. Slowly rotate and withdraw the socket assembly from the housing. Note the orientation and placement of the metal shield and the black plastic insulating cover.

With reaction chamber now assembled, replace and reconnect it in reversed re- moval sequence. Orient as noted in step 2.

3. Carefully unplug the photomultiplier tube from the socket.
4. Plug a new tube into the socket.

Rosemount Analytical Inc. A Division of Emerson Process Management Routine Servicing 5-5

Page 46 of the Emerson Controller (951C) user manual — NOTE:
Page 46 · click to enlarge
Page 46
5. Orient the metal shield and black plastic insulator as noted in step 2.

a. Lamp/Housing Removal

To remove the lamp and housing, do the follow:

6. Carefully rotate and insert the tube, shield and cover into the housing. Orient as noted in step 1.
1. Disconnect power from the instrument.

5-4 OZONE GENERATION SYSTEM

2. Release pressure from SAMPLE and AIR supplies.

This system consists of the ultraviolet lamp, lamp housing, and power supply. Refer to Figure 6-1 for location and details of mount- ing.

3. Disconnect the air supply tubing from the front of the housing.

WARNING TOXIC CHEMICAL HAZARD

4. Disconnect the ozone tube leading to the reaction chamber.

The ozone generator lamp contains mercury. Lamp breakage could result in mercury exposure. Mercury is highly toxic if absorbed through skin or ingested, or if vapors are inhaled. Handle lamp assembly with extreme care.

5. Disconnect the power cable from the Power Supply.
6. Uncouple the two Velcro straps which secure the housing to power supply.

If lamp is broken, avoid skin contact and inhalation in the area of the lamp or the mercury spill. Immediately clean up and dispose of the mercury spill and lamp residue as follows:

7. Lift the housing from the analyzer.

b. UV Lamp Replacement

To replace the lamp, do the following:

  • Wearing rubber gloves and goggles, collect all droplets of mercury by means of a suction pump and aspirator bottle with long capillary tube. Alternatively, a commercially available mercury spill clean-up kit, such as J. T. Baker product No. 4439-01, is recommended.
1. Unscrew and remove end cap.
2. Unscrew aluminum outer lamp housing tube from lamp base, using care not to hit or touch lamp assembly.

NOTE:

Do not touch lamp. Fingerprints may cause a decrease in lamp output.

  • Carefully sweep any remaining mercury and lamp debris into a dust pan. Carefully transfer all mercury, lamp residue and debris into a plastic bottle which can be tightly capped. Label and return to hazardous material reclamation center.
3. Replace O-ring in lamp base with new O-ring supplied in kit.
4. Insert replacement lamp assembly using care not to hit or touch lamp housing.

Do not place in trash, incinerate or flush down sewer. Cover any fine droplets of mercury in non-accessible crevices with calcium polysulfide and sulfur dust.

5. Insert new O-ring into new end cap. Screw end cap onto end of lamp housing.

Replace the lamp and housing by revers- ing the steps in this section.

5-6 Routine Servicing Rosemount Analytical Inc. A Division of Emerson Process Management

Page 47 of the Emerson Controller (951C) user manual — 5-5 CONVERTER ASSEMBLY
Page 47 · click to enlarge
Page 47

c. Power Supply Removal

than the outlet tube. Further, the outlet tube and the sample capillary (P/N 615989) connect to the same stainless steel tee.

To remove the Power Supply, do the fol- lowing:

Refer to Figure 6-1.

3. Release the assembly and disconnect the heater and sensor connectors from the temperature control board.
1. Remove the lamp and housing as in Section 5-4b.
4. Remove the lacing from the heater blanket, and remove the converter tube. Note the position of the temperature sensor and its leads as the aluminum foil is unwrapped.
2. Disconnect the power lead from the Power Supply Board.
3. Remove the two screws which secure the Power Supply to the bottom plate of the analyzer.
5. Replace the defective part and reassemble. The temperature sensor should contact the converter tube with the top of the sensor at the midpoint of the converter. Route sensor leads axially to the outer end.
4. Lift the Power Supply from the analyzer.
5. Replace the Power Supply by reversing the order of the steps in this section.
6. Condition the converter as described in Sections 3-4 and 3-5.

5-5 CONVERTER ASSEMBLY

5-6 SERVICING ELECTRONIC CIRCUITRY

To check the heater blanket, verify the conti- nuity of the heater coil.

For troubleshooting the electronic system, re- fer to Section 4.0 and the appropriate pictorial diagrams at the back of the manual. The elec- tronic system utilizes printed circuit boards with solid-state components. After a malfunc- tion is traced to a particular board, the rec- ommended procedure is to return it to the factory for repair.

To check the temperature sensor, refer to Section 3-4 and measure its resistance when instrument power is off (should be about 440 ohms) and when instrument power is on (should range from 800 to 1,000 ohms). See Table 3-1.

To remove the glass converter tube (see Figure 6-4):

1. Carefully disconnect the blue silicon connectors from the ends of the inlet and outlet tubes.
2. The inlet tube is partially filled with glass wool and has a larger inside diameter

Rosemount Analytical Inc. A Division of Emerson Process Management Routine Servicing 5-7

Page 48 of the Emerson Controller (951C) user manual — SECTION 4
Page 48 · click to enlarge
Page 48

5-8 Routine Servicing Rosemount Analytical Inc. A Division of Emerson Process Management

Page 49 of the Emerson Controller (951C) user manual — 6-1 MATRIX
Page 49 · click to enlarge
Page 49

SECTION 6 REPLACEMENT PARTS

Emerson Controller (951C) — SECTION 6 REPLACEMENT PARTS diagram (page 49)
Fig. 18

Tampering with or unauthorized substitution of components may adversely affect safety of this product. Use only factory-approved components for repair.

6-1 MATRIX

To identify the configuration of an analyzer, locate the analyzer name-rating plate. The sales matrix identifier number appears on the analyzer name-rating plate.

Each analyzer is configured per the customer sales order. Below is the 951C sales matrix which lists the various configurations avail- able.

Emerson Controller (951C) — 6-1 MATRIX diagram (page 49)
Fig. 19

951C Process Chemiluminescence NOx Analyzer (19" Rack Mount) (951C)

01 Low Ranges0-10, 0-25, 0-100, 0-250 ppm NOx
02 High Ranges0-100, 0-250, 0-1000, 0-2500 ppm NOx
04 Mid Ranges0-20, 0-50, 0-200, 0-500 ppm NOx

02 User controlled sample flow (no sample restrictor included)

Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 6-1

Page 50 of the Emerson Controller (951C) user manual — 6-3 REPLACEMENT PARTS
Page 50 · click to enlarge
Page 50

6-2 CIRCUIT BOARD REPLACEMENT POLICY

a. Common Parts

In most situations involving a malfunction of a circuit board, it is more practical to replace the board than to attempt isolation and replace- ment of the individual component. The cost of test and replacement will exceed the cost of a rebuilt assembly from the factory.

Refer to Figure 6-1.

655519 Air Restrictor Fitting 657091 Capacitor Assembly 655166 Capillary, Bypass 655589 Capillary, Sample Hi 623719 Capillary, Sample Lo 654068 Temperature Control Assembly 654070 Converter Assembly 655303 Exhaust Fan 654052 Fan Assembly 898587 Fuse 3.15 A 902413 Fuse 6.25 A 654390 I/O Assembly 652173 Ozone Generator 658156 Ozone Generator UV Lamp Replacement Kit 655129 Ozone Generator Power Supply 654062 Photomultiplier Assembly 655332 Power Supply Assembly 662273 Pressure Switch 623936 Sample Flow Restrictor 644055 Sample Pressure Gauge 815187 Sample Regulator 622917 Sensor, Temperature 6A00337G01/02/03 Signal/Control Board 654878 Transformer/Inductor Assembly

The following list does not include individual electronic components. If circumstances ne- cessitate replacement of an individual compo- nent which can be identified by inspection or from the schematic diagrams, obtain the re- placement component from a local source of supply.

6-3 REPLACEMENT PARTS

The following parts are recommended for rou- tine maintenance and troubleshooting of the Model 951C NOX Analyzer. If the trouble- shooting procedures do not resolve the prob- lem, contact your local Rosemount Analytical service office. A list of Rosemount Analytical Service Centers is located in Section 7.

Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 6-3

Page 51 of the Emerson Controller (951C) user manual — SECTION 4
Page 51 · click to enlarge
Page 51

Photomultiplier Assembly 654062 (See Figure 6-2)

Emerson Controller (951C) — SECTION 4 diagram (page 51)
Fig. 20

Case Heater Temperature Control Assembly 654068 (See Figure 6-4)

Sample Regulator 815187 Sample Pressure Gauge 644055

Figure 6-1. Major Assemblies of the Model 951C

Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 6-3

Page 52 of the Emerson Controller (951C) user manual — SECTION 4
Page 52 · click to enlarge
Page 52

b. Photomultiplier Assembly 654062

001522 O-Ring 008423 O-Ring, Photomultiplier 655168 Photomultiplier Tube 654381 Reaction Chamber 654086 Socket Assembly 639722 Thermal Shield

Refer to Figure 6-2.

654943 Housing 649541 Insulating Washer 636318 Magnetic Shield 630916 Magnetic Shield

Emerson Controller (951C) — SECTION 4 diagram (page 52)
Fig. 21

Photomultiplier Tube and Magnetic Shield (PN 630916) to be Flush

Figure 6-2. Photomultiplier Housing Assembly

6-4 Replacement Parts Rosemount Analytical Inc. A Division of Emerson Process Management

Page 53 of the Emerson Controller (951C) user manual — SECTION 4
Page 53 · click to enlarge
Page 53

c. Converter Assembly 654070

Refer to Figure 6-3.

632784 Connector, Blue Silicone 657127 Heater 632782 Temperature Sensor 632795 Tube

Emerson Controller (951C) — SECTION 4 diagram (page 53)
Fig. 22

Figure 6-3. Converter Assembly

Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 6-5

Page 54 of the Emerson Controller (951C) user manual — SECTION 4
Page 54 · click to enlarge
Page 54

d. Temperature Control Assembly 654068

Refer to Figure 6-4.

622733 Fan 622732 Heater 655335 Temperature Control Board 900492 Thermal Fuse

Emerson Controller (951C) — SECTION 4 diagram (page 54)
Fig. 23

Figure 6-4. Case Heater Temperature Control Assembly

6-6 Replacement Parts Rosemount Analytical Inc. A Division of Emerson Process Management

Page 55 of the Emerson Controller (951C) user manual — SECTION 7 RETURN OF MATERIAL
Page 55 · click to enlarge
Page 55

SECTION 7 RETURN OF MATERIAL

7-1 RETURN OF MATERIAL If factory repair of defective equipment is re- quired, proceed as follows: 1. Secure a return authorization from a Rosemount Analytical Inc. Sales Office or Representative before returning the equipment. Equipment must be returned with complete identification in accordance with Rosemount instructions or it will not be accepted. Rosemount CSC (Customer Service Cen- ter) will provide the shipping address for your instrument. In no event will Rosemount be responsi- ble for equipment returned without proper authorization and identification. 2. Carefully pack the defective unit in a sturdy box with sufficient shock absorbing material to ensure no additional damage occurs during shipping. 3. In a cover letter, describe completely:

If warranty service is expected, the defective unit will be carefully inspected and tested at the factory. If the failure was due to the condi- tions listed in the standard Rosemount war- ranty, the defective unit will be repaired or replaced at Rosemount’s option, and an oper- ating unit will be returned to the customer in accordance with the shipping instructions fur- nished in the cover letter. For equipment no longer under warranty, the equipment will be repaired at the factory and returned as directed by the purchase order and shipping instructions.

7-2 CUSTOMER SERVICE For order administration, replacement Parts, application assistance, on-site or factory re- pair, service or maintenance contract informa- tion, contact:

  • The symptoms that determined the equipment is faulty.
  • The environment in which the equipment was operating (housing, weather, vibration, dust, etc.).
  • Site from where the equipment was removed.
  • Whether warranty or non-warranty service is expected.
  • Complete shipping instructions for the return of the equipment. 4. Enclose a cover letter and purchase order and ship the defective equipment according to instructions provided in the Rosemount Return Authorization, prepaid, to the address provided by Rosemount CSC.

Rosemount Analytical Inc.

7-3 TRAINING A comprehensive Factory Training Program of operator and service classes is available. For a copy of the Current Operator and Service Training Schedule contact :

Rosemount Analytical Inc.

Rosemount Analytical Inc.

7-2 Replacement Parts Rosemount Analytical Inc. A Division of Emerson Process Management

Show 9 more labels / short text on this page
Process Analytic Division · Customer Service Center · 1-800-433-6076 · Process Analytic Division · Customer Service Center · 1-800-433-6076 · Process Analytic Division · Customer Service Center · 1-800-433-6076
Page 56 of the Emerson Controller (951C) user manual — SECTION 4
Page 56 · click to enlarge
Page 56

7-2 Return of Material Rosemount Analytical Inc. A Division of Emerson Process Management

Page 57 of the Emerson Controller (951C) user manual — R Range2 Select jumpers, 3-3
Page 57 · click to enlarge
Page 57

SECTION 8

INDEX

establish correct pressure of zero gas, 3-4 Excessive Photomultiplier Dark Current, 5-2

2 230 VAC conversion, 2-3

F flow rate, measuring, 5-3 flow rate, subnormal, 5-3

A Air (U.S.P. Breathing Grade), 2-5 AIR inlet

adequate air pressure, 3-3 setpoint pressure, 3-3 analyzer

I Inability to obtain a flow of one liter per minute at the

EXHAUST outlet, 3-4 internal SAMPLE pressure, 3-1

installation location, 2-3 mount near sample, 2-3 operating temperature, 2-3 warm-up, 7, 3-3

O output connections

C Contamination of Reaction Chamber or Sample Flow

routing, 2-3 ozonator air

supply pressure, 7 ozone generation system, description, 5-6 ozone generation system, power supply removal, 5-

glass converter tube, 5-7 heater blanket, 5-7 measuring efficiency, 3-9 optimizing operating temperature, 3-7 temperature check, 3-7 temperature sensor, 5-7 converter temperature

7 ozone generation system, removal, 5-6 ozone generation system, UV lamp replacement, 5-7 ozone lamp

output check, 5-2 OZONE pressure

adjustment, 3-7 recommended periodic check, 3-7 current output

determination, 3-1 nominal, 3-1

adjustment, 2-3 Current recorder connections

P photomultiplier tube, 1-1, 4-1, 5-1, 5-2, 5-3, 5-4, 5-6 photomultiplier tube, description, 5-5 potentiometric recorder connections

cable specifications, 2-3 hookup, 2-3 routing, 2-3 cylinders, air/gas

cable specifications, 2-3 hookup, 2-3 routing, 2-3 Power connections

D display, 3-1 DP SELECT jumpers, 3-3

cable type, 2-3 hookup, 2-2 routing, 2-3 PPM RANGE Switch, 3-3

E Establish correct flow of upscale standard gas, 3-4 Establish correct pressure for air, 3-4 Establish correct pressure of sample gas, 3-4

R Range2 Select jumpers, 3-3

Rosemount Analytical Inc. A Division of Emerson Process Management Index 8-1

adjustment, 3-1 · System, 5-2 converter · location, 2-3

Page 58 of the Emerson Controller (951C) user manual — T thermoelectric cooler, 5-4
Page 58 · click to enlarge
Page 58

reaction chamber, 4-1, 4-2, 5-1, 5-2, 5-3, 5-4, 5-5, 5-

U Upscale Calibration, 8

6 reaction chamber, alternate cleaning procedure, 5-5 reaction chamber, standard cleaning procedure, 5-5 Return Authorization, 7-2

PPM RANGE switch, 3-6 RANGE1 control, 3-6 sensitivity, 3-6 supply gas, 3-6

S sample gas

Z Zero Calibration, 7

establish correct pressure, 7 supply pressure, 7 shutdown, 3-6 Span Gas, 2-5 supply pressure, 2-5

PPM RANGE switch, 3-6 RANGE1 control, 3-6 ZERO control, 3-6 zero display reading, inability to obtain, 5-2 zero gas

T thermoelectric cooler, 5-4

establish correct pressure, 7 supply pressure, 7

8-4 Index Rosemount Analytical Inc. A Division of Emerson Process Management

Page 59 of the Emerson Controller (951C) user manual — LIST OF DRAWINGS
Page 59 · click to enlarge
Page 59

LIST OF DRAWINGS

Emerson Controller (951C) — LIST OF DRAWINGS diagram (page 59)
Fig. 24

Drawing 1-1. Panel Cutout / Installation Drawing

Rosemount Analytical Inc. A Division of Emerson Process Management List of Drawings

Page 60 of the Emerson Controller (951C) user manual — SECTION 4
Page 60 · click to enlarge
Page 60
Emerson Controller (951C) — SECTION 4 diagram (page 60)
Fig. 25

Drawing 1-2 Flow Diagram, Lo/Mid Range

Rosemount Analytical Inc. A Division of Emerson Process Management

List of Drawings

Page 61 of the Emerson Controller (951C) user manual — SECTION 4
Page 61 · click to enlarge
Page 61
Emerson Controller (951C) — SECTION 4 diagram (page 61)
Fig. 26

Drawing 1-3 Flow Diagram, Hi Range

Rosemount Analytical Inc. A Division of Emerson Process Management List of Drawings

Page 62 of the Emerson Controller (951C) user manual — WARRANTY
Page 62 · click to enlarge
Page 62

WARRANTY

Goods and part(s) (excluding consumables) manufactured by Seller are warranted to be free from defects in workmanship and material under normal use and service for a period of twelve (12) months from the date of shipment by Seller. Consumables, glass electrodes, membranes, liquid junctions, electrolyte, o-rings, etc., are warranted to be free from defects in workmanship and material under normal use and service for a period of ninety (90) days from date of shipment by Seller. Goods, part(s) and consumables proven by Seller to be defective in workmanship and/or material shall be replaced or repaired, free of charge, F.O.B. Seller's factory provided that the goods, part(s) or consumables are returned to Seller's designated factory, transportation charges prepaid, within the twelve (12) month period of warranty in the case of goods and part(s), and in the case of consumables, within the ninety (90) day period of warranty. This warranty shall be in effect for replacement or repaired goods, part(s) and the remaining portion of the ninety (90) day warranty in the case of consumables. A defect in goods, part(s) and consumables of the commercial unit shall not operate to condemn such commercial unit when such goods, part(s) and consumables are capable of being renewed, repaired or replaced.

The Seller shall not be liable to the Buyer, or to any other person, for the loss or damage directly or indirectly, arising from the use of the equipment or goods, from breach of any warranty, or from any other cause. All other warranties, expressed or implied are hereby excluded.

IN CONSIDERATION OF THE HEREIN STATED PURCHASE PRICE OF THE GOODS, SELLER GRANTS ONLY THE ABOVE STATED EXPRESS WARRANTY. NO OTHER WARRANTIES ARE GRANTED INCLUDING, BUT NOT LIMITED TO, EXPRESS AND IMPLIED WARRANTIES OR MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.

Limitations of Remedy. SELLER SHALL NOT BE LIABLE FOR DAMAGES CAUSED BY DELAY IN PERFORMANCE. THE SOLE AND EXCLUSIVE REMEDY FOR BREACH OF WARRANTY SHALL BE LIMITED TO REPAIR OR REPLACEMENT UNDER THE STANDARD WARRANTY CLAUSE. IN NO CASE, REGARDLESS OF THE FORM OF THE CAUSE OF ACTION, SHALL SELLER'S LIABILITY EXCEED THE PRICE TO BUYER OF THE SPECIFIC GOODS MANUFACTURED BY SELLER GIVING RISE TO THE CAUSE OF ACTION. BUYER AGREES THAT IN NO EVENT SHALL SELLER'S LIABILITY EXTEND TO INCLUDE INCIDENTAL OR CONSEQUENTIAL DAMAGES. CONSEQUENTIAL DAMAGES SHALL INCLUDE, BUT ARE NOT LIMITED TO, LOSS OF ANTICIPATED PROFITS, LOSS OF USE, LOSS OF REVENUE, COST OF CAPITAL AND DAMAGE OR LOSS OF OTHER PROPERTY OR EQUIPMENT. IN NO EVENT SHALL SELLER BE OBLIGATED TO INDEMNIFY BUYER IN ANY MANNER NOR SHALL SELLER BE LIABLE FOR PROPERTY DAMAGE AND/OR THIRD PARTY CLAIMS COVERED BY UMBRELLA INSURANCE AND/OR INDEMNITY COVERAGE PROVIDED TO BUYER, ITS ASSIGNS, AND EACH SUCCESSOR INTEREST TO THE GOODS PROVIDED HEREUNDER.

Force Majeure. Seller shall not be liable for failure to perform due to labor strikes or acts beyond Seller's direct control.

Rosemount Analytical Inc. A Division of Emerson Process Management

Page 63 of the Emerson Controller (951C) user manual — SECTION 4
Page 63 · click to enlarge
Page 63
Page 63 of the Emerson Controller (951C) user manual — SECTION 4
Page 64 of the Emerson Controller (951C) user manual — T 44 1243 863121 F 44 1243 845354
Page 64 · click to enlarge
Page 64

WORLD HEADQUARTERS AND AMERICAS

ROSEMOUNT ANALYTICAL EUROPE Emerson Process Management GmbH & Co. OHG

LATIN AMERICA

Emerson Process Management Rosemount Analytical

Emerson Process Management Rosemount Analytical Inc.

Industriestrasse 1 63594 Hasselroth

11100 Brittmoore Park Road Houston, TX 77041 USA T 713.467.6000 F 713.827.3328F 49 6055 884209

6565 P Davis Industrial Parkway Solon, OH 44139 USA

Germany T 49 6055 884 0 F 49 6055 884209

T 440.914.1261 Toll Free in US and Canada 800.433.6076 F 440.914.1271

e-mailgas.csc@ Emerson.com www.raihome.com National Response Center 800.654.7768

ASIA-PACIFIC Emerson Process Management

EUROPE Emerson Process Management Shared Services Limited

MIDDLE EAST AND AFRICA Emerson Process Management

Asia Pacific Private Limited

EPM Building P.O. Box 17033 Jebel Ali Free Zone

1 Pandan Crescent Singapore 128461

Heath Place Bognor Regis

Republic of Singapore T 65 6 777 8211

West Sussex PO22 9SH England

Dubai, United Arab Emirates T 971 4 8835235 F 971 4 8835312

F 65 6 777 0947 e-mail: analytical@ap.emersonprocess.com

T 44 1243 863121 F 44 1243 845354

© Rosemount Analytical Inc. 2009

Sources

References & links

Pages referenced in this manual — manufacturer, support and downloads.

Ask about the Emerson Controller (951C)

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the Emerson Controller (951C)

Be the first to ask — other owners or our team will answer.

Original document

Emerson Controller (951C) manual PDF

64 pages·1.8 MB·Controller·PDF
Click any page above to open it here
Page 1 / …
100%
Loading pages…