WARNINGS
READ THIS PAGE BEFORE PROCEEDING! Rosemount Analytical designs, manufactures, and tests its prod- ucts to meet many national and international standards. Because these instruments are sophisticated technical products, you must properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions must be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount Analytical products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.
Making cable connections to and servicing this instru- ment require access to shock hazard level voltages which can cause death or serious injury, therefore, disconnect all hazardous voltage before accessing the electronics.
Relay contacts made to separate power sources must be disconnected before servicing.
Electrical installation must be in accordance with the National Electrical Code (ANSI/NFPA-70) and/or any other applicable national or local codes.
Unused cable conduit entries must be securely sealed by non-flammable closures to provide enclo- sure integrity in compliance with personal safety and environmental protection requirements. Use NEMA 4X or IP65 conduit plugs supplied with the instrument to maintain the ingress protection rating (IP65).
For safety and proper performance this instrument must be connected to a properly grounded three- wire power source.
Proper relay use and configuration is the responsi- bility of the user. No external connection to the instrument of more than 60VDC or 43V peak allowed with the exception of power and relay termi- nals. Any violation will impair the safety protection provided.
Do not operate this instrument without front cover secured. Refer installation, operation and servicing to qualified personnel.
This product is not intended for use in the res- idential, commercial or light industrial environ- ment per certification to EN50081-2.



MODEL 54e pH/ORP SPECIFICATIONS
NEMA 4X (IP65) 144 X 144 X 131 mm, DIN size (5.7 X 5.7 X 5.2 in.) Front Panel: Membrane keyboard with tactile
contacts, SPST, normally open
Resistive Inductive 28 Vdc 5.0 Amps 3.0 Amps 115 Vac 5.0 Amps 3.0 Amps 230 Vac 5.0 Amps 1.5 Amps
feedback and user selectable security. Light gray, blue and white overlay Light gray enclosure, dark gray bezel
blue on gray-green The display contrast is compensated for ambient temperature.
Electrical Classification:
Class I, Division 2, Groups A, B, C, & D. T5 Ta=50°C. Dust ignition proof: Class II, Division 1, Groups E, F, & G; Class III.
150 mA - Groups A & B; 400 mA - Group C; 540 mA - Group D Power:
Automatic or Manual; –15 to 120°C (5 to 248°F)
glass impedance, or reference impedance. Output 2: pH, ORP, temperature, glass impedance, or reference impedance. Each output is galvanically isolated, 0-20 mA or 4-20 mA into 500 ohms maximum load at 24 Vdc or 115/230 Vac or 500 ohms maximum load at 100/200 Vac. Output 1 includes digital signal 4-20 mA superimposed HART (Code -09 only).
Pt 100 or Pt 1000 RTD
Alarms:
Relay 1 - Process, Interval, or Time Proportional
Control (TPC requires code -20) Relay 2 - Process, Interval, or Time Proportional Control (TPC requires code -20) Relay 3 - Process, Interval, or Time Proportional Control (TPC requires code -20) Relay 4 - Sensor/analyzer and process fault alarm Each relay has a dedicated LED on the front panel.
Sira MC070111/00

MODEL 54e pH/ORP INSTALLATION
Pipe Mounting:
Position the Model 54e pH/ORP controller to minimize the effects of temperature extremes and to avoid vibration and shock. Locate the controller away from your chemical process to protect it from moisture and fumes.
Panel Mounting:
The controller is designed to fit into a 5.43 x 5.43 inch (DIN stan- dard 144x144 mm) panel cutout (shown below). Installation requires both front and rear access.
Select an installation site that is more than 2 ft from high voltage conduit, has easy access for operating personnel, and is not exposed to direct sunlight.
Inspect the exterior of the shipping container for any damage. Open the container and inspect the controller and related hard- ware for missing or damaged parts.
If there is evidence of damage, notify the carrier immediately. If parts are missing, contact Rosemount Analytical customer sup- port.

The Model 54e pH/ORP controller may be supplied with a mount- ing bracket accessory. If you use the mounting bracket on wall or pipe installations, avoid mounting on pipes which vibrate or are close to the process. The bracket may be modified to mount the controller on I-beams or other rigid members. You can also fabri- cate your own bracket or panel mount the controller using the bracket as an example.
Wall or Surface Mounting:


PIPE MOUNTING
Mounting the Controller

MODEL 54e pH/ORP MENUS

All menus shown. Some menus will change or not appear with different settings.

MODEL 54e pH/ORP QUICK START
Use the ↓or key to move the cursor down or up and the → or ←to move the cursor right or left.
Press ENTER (F4) to move to the next menu, SAVE (F4) to store a setting, EDIT (F4) to change a setting, and EXIT (F1) to exit without saving changes.
Press any button. The main menu will appear with CAL- IBRATE highlighted. Press ENTER.
BUFFER CALIBRATION will be highlighted. Press ENTER.
Place the sensor in the first buffer and press CONT. Wait until the screen reads "Buf 1 done."
Press any button. he main menu will appear with CALI- BRATE highlighted.
Use the arrow buttons to select the correct buffer. Press CONT.
Use the arrow buttons to select Program. Press ENTER.
Rinse and dry the sensor.
Use the arrow buttons to select Configure. Press ENTER.
Place the sensor in the second buffer and press CONT. Wait until the screen reads "Buf 2 done."
DISPLAY will be highlighted. Press ENTER.
Use the arrow buttons to select the correct buffer. Press CONT.
MEASURE will be highlighted. Press ENTER.
Press EXIT until the main display appears.
Use the arrow buttons to select REDOX or ORP. Press SAVE.
Press EXIT until the main display appears.
STEP 2 - OUTPUT SETPOINTS
After the sensor is installed, take a grab sample as close to the sensor as possible.
Press any button. The main menu will appear with CAL- IBRATE highlighted.
Use a calibrated pH instrument with automatic tempera- ture compensation to determine the pH of the grab sam- ple.
Use the arrow buttons to select Program. Press ENTER.
Press any button. The main menu will appear with CAL- IBRATE highlighted. Press ENTER.
Use the arrow buttons to select OUTPUT SETPOINTS. Press ENTER.
Use the arrow buttons to select STANDARDIZE. Press ENTER.
OUTPUT 1 SETPOINTS will be highlighted. Press ENTER. Press CONT.
Press EDIT. Use the arrow buttons to enter the pH value of the sample. Press SAVE.
4 mA will be highlighted. Press EDIT. Use the arrow but- tons to enter the desired value. Press SAVE.
Press EXIT until the main display appears.
Use the arrow buttons to select 20 MA. Press EDIT. Use the arrow buttons to enter the desired value. Press SAVE. Press EXIT.
Use the arrow buttons to select OUTPUT 2 SET- POINTS. Press ENTER. Press CONT.
Enter the 4 mA and 20 mA values as above.
Press EXIT until the main display appears.

MODEL 54e pH/ORP WIRING

Sensor Wiring Diagram



Power Input and Relay Output Wiring for Model 54e pH/ORP

The right people, the right answers, right now. ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE
Specifications subject to change without notice.
Credit Cards for U.S. Purchases Only.
2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250
© Rosemount Analytical Inc. 2010
Emerson Process Management