Emerson 54e

8 pages·506 KB·Marine Instruments·PDF·ENDEFRITNLJA
Emerson 54e
Table of Contents 24 sections Instruction Sheet PN 51A-54epH/rev J October 2010 Model 54e pH/ORP · ® Analyzer/Controller · pH/ORP HART · For additional information, please visit our website · … ShowHide
Page 1 of the Emerson 54e user manual — WARNING
Page 1 · click to enlarge
Page 1

Instruction Sheet PN 51A-54epH/rev J October 2010 Model 54e pH/ORP

® Analyzer/Controller

pH/ORP HART

For additional information, please visit our website

at www.emersonprocess.com/raihome/liquid/.

ESSENTIAL INSTRUCTIONS

WARNINGS

READ THIS PAGE BEFORE PROCEEDING! Rosemount Analytical designs, manufactures, and tests its prod- ucts to meet many national and international standards. Because these instruments are sophisticated technical products, you must properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions must be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount Analytical products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.

RISK OF ELECTRICAL SHOCK

Making cable connections to and servicing this instru- ment require access to shock hazard level voltages which can cause death or serious injury, therefore, disconnect all hazardous voltage before accessing the electronics.

Relay contacts made to separate power sources must be disconnected before servicing.

Electrical installation must be in accordance with the National Electrical Code (ANSI/NFPA-70) and/or any other applicable national or local codes.

  • Read all instructions prior to installing, operating, and servicing the product. If this Instruction Manual is not the correct manual, telephone 1-949-757-8500 and the requested manual will be provided. Save this Instruction Manual for future reference.

Unused cable conduit entries must be securely sealed by non-flammable closures to provide enclo- sure integrity in compliance with personal safety and environmental protection requirements. Use NEMA 4X or IP65 conduit plugs supplied with the instrument to maintain the ingress protection rating (IP65).

  • If you do not understand any of the instructions, contact your Rosemount representative for clarification.
  • Follow all warnings, cautions, and instructions marked on and supplied with the product.

For safety and proper performance this instrument must be connected to a properly grounded three- wire power source.

  • Inform and educate your personnel in the proper installation, operation, and maintenance of the product.

Proper relay use and configuration is the responsi- bility of the user. No external connection to the instrument of more than 60VDC or 43V peak allowed with the exception of power and relay termi- nals. Any violation will impair the safety protection provided.

  • Install your equipment as specified in the Installation Instructions of the appropriate Instruction Manual and per applicable local and national codes. Connect all products to the proper electrical and pressure sources.
  • To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the product.

Do not operate this instrument without front cover secured. Refer installation, operation and servicing to qualified personnel.

  • When replacement parts are required, ensure that qualified people use replacement parts specified by Rosemount. Unauthorized parts and procedures can affect the product’s performance and place the safe operation of your process at risk. Look alike substitutions may result in fire, electrical hazards, or improper operation.

WARNING

This product is not intended for use in the res- idential, commercial or light industrial environ- ment per certification to EN50081-2.

  • Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being performed by qualified persons, to prevent electrical shock and personal injury.
Page 2 of the Emerson 54e user manual — Instruction Sheet PN 51A-54epH/rev J October 2010 Model 54e pH/ORP
Page 2 · click to enlarge
Page 2
Emerson 54e — Instruction Sheet PN 51A-54epH/rev J October 2010 Model 54e pH/ORP diagram (page 2)
Fig. 1 · Instruction Sheet PN 51A-54epH/rev J October 2010 Mo…
Page 3 of the Emerson 54e user manual — RFI/EMI: EN-61326
Page 3 · click to enlarge
Page 3

MODEL 54e pH/ORP SPECIFICATIONS

PHYSICAL SPECIFICATIONS – GENERAL

Relay ContactsRelays 1-3: Epoxy sealed form A
EnclosureEpoxy-painted aluminum

NEMA 4X (IP65) 144 X 144 X 131 mm, DIN size (5.7 X 5.7 X 5.2 in.) Front Panel: Membrane keyboard with tactile

contacts, SPST, normally open

Relay 4Epoxy sealed form C, SPDT

Resistive Inductive 28 Vdc 5.0 Amps 3.0 Amps 115 Vac 5.0 Amps 3.0 Amps 230 Vac 5.0 Amps 1.5 Amps

feedback and user selectable security. Light gray, blue and white overlay Light gray enclosure, dark gray bezel

DisplayBack-lit dot matrix LCD (7.0 x 3.5 cm),

blue on gray-green The display contrast is compensated for ambient temperature.

Weight/Shipping Weight1.8 kg/2.3 kg (4 lb/5 lb)
Process Variable Character Height16mm (0.6 in.)

Electrical Classification:

INSTRUMENT SPECIFICATIONS @ 25°C — pH

Class I, Division 2, Groups A, B, C, & D. T5 Ta=50°C. Dust ignition proof: Class II, Division 1, Groups E, F, & G; Class III.

Measurement Range0 to 14 pH
Output Scale ExpansionZero suppression: up to 13

CSA-LR34186:

pH units SpanAny pH from 1 to 14
Max. relay contact rating28 Vdc; 110 Vac; 230 Vac; 6 amps resistive
FMMax. relay contact rating: 28 Vdc resistive
Accuracy± 0.01 pH

150 mA - Groups A & B; 400 mA - Group C; 540 mA - Group D Power:

Repeatability± 0.01 pH
Stability± 0.01 pH/month, non-cumulative
Temperature CoefficientInput: ± 0.003 pH/°C
Code -01100 - 127 VAC, 50/60 Hz ± 6%, 6.0 W
Output± 0.006 pH/°C
200. 253 VAC, 50/60 Hz ± 6%, 6.0 W Code -02: 20 - 30 VDC, 6.0 W
Temperature CompensationPt 100 or Pt 1000 RTD,

Automatic or Manual; –15 to 120°C (5 to 248°F)

Current OutputsOutput 1: pH, ORP, temperature,

glass impedance, or reference impedance. Output 2: pH, ORP, temperature, glass impedance, or reference impedance. Each output is galvanically isolated, 0-20 mA or 4-20 mA into 500 ohms maximum load at 24 Vdc or 115/230 Vac or 500 ohms maximum load at 100/200 Vac. Output 1 includes digital signal 4-20 mA superimposed HART (Code -09 only).

INSTRUMENT SPECIFICATIONS @ 25°C — ORP

Measurement Range–1400 to +1400 mV
Output Scale ExpansionZero suppression: up to
±1300 mV SpanAny ORP range from 100 to 2800 mV

RFI/EMI: EN-61326

Accuracy± 1.0 mV
Repeatability± 1.0 mV
LVD (Code -01 only)EN-61010-1
Stability± 1.0 mV/month, non-cumulative
Ambient Temperature0 to 50°C (32 to 122°F)
Temperature CoefficientInput: ± 0.2 mV/°C
Output± 0.4 mV/°C
Temperature Measurement–15 to 120°C (5 to 248°F)
Relative Humidity95%, non condensing

Pt 100 or Pt 1000 RTD

Alarms:

Relay 1 - Process, Interval, or Time Proportional

Control (TPC requires code -20) Relay 2 - Process, Interval, or Time Proportional Control (TPC requires code -20) Relay 3 - Process, Interval, or Time Proportional Control (TPC requires code -20) Relay 4 - Sensor/analyzer and process fault alarm Each relay has a dedicated LED on the front panel.

Sira MC070111/00

Page 4 of the Emerson 54e user manual — WALL MOUNTING
Page 4 · click to enlarge
Page 4

MODEL 54e pH/ORP INSTALLATION

Pipe Mounting:

1. Attach the mounting bracket to the rear of the controller and tighten the four screws.
2. Place supplied U bolts around the mounting pipe and through the pipe mounting bracket and mounting bracket. Tighten the U bolt nuts until the controller is securely mounted to the pipe.

LOCATING THE CONTROLLER

Position the Model 54e pH/ORP controller to minimize the effects of temperature extremes and to avoid vibration and shock. Locate the controller away from your chemical process to protect it from moisture and fumes.

Panel Mounting:

The controller is designed to fit into a 5.43 x 5.43 inch (DIN stan- dard 144x144 mm) panel cutout (shown below). Installation requires both front and rear access.

Select an installation site that is more than 2 ft from high voltage conduit, has easy access for operating personnel, and is not exposed to direct sunlight.

1. Install the controller as shown below. Insert the instrument enclosure through the front of the panel cutout and align the panel mounting brackets as shown.
2. Insert two mounting bracket screws through each of the two mounting brackets and into the tapped holes in the rear of the controller enclosure and tighten each screw.

UNPACKING AND INSPECTION

Inspect the exterior of the shipping container for any damage. Open the container and inspect the controller and related hard- ware for missing or damaged parts.

3. Insert four panel mounting screws through each hole in the mounting brackets. Tighten each screw until the mounting bracket holds controller firmly in place. To avoid damaging the controller mounting brackets, do not use excessive force.

If there is evidence of damage, notify the carrier immediately. If parts are missing, contact Rosemount Analytical customer sup- port.

Emerson 54e — UNPACKING AND INSPECTION diagram (page 4)
Fig. 2

MECHANICAL INSTALLATION

The Model 54e pH/ORP controller may be supplied with a mount- ing bracket accessory. If you use the mounting bracket on wall or pipe installations, avoid mounting on pipes which vibrate or are close to the process. The bracket may be modified to mount the controller on I-beams or other rigid members. You can also fabri- cate your own bracket or panel mount the controller using the bracket as an example.

Wall or Surface Mounting:

1. Mount the bracket to the controller using the supplied four screws.
2. Mount controller mounting bracket to wall using any appropriate fastener such as screws, bolts, etc

PANEL MOUNTING

Emerson 54e — PANEL MOUNTING diagram (page 4)
Fig. 3
Emerson 54e — PANEL MOUNTING diagram (page 4)
Fig. 4

WALL MOUNTING

PIPE MOUNTING

Mounting the Controller

Page 5 of the Emerson 54e user manual — LOCATING THE CONTROLLER
Page 5 · click to enlarge
Page 5

MODEL 54e pH/ORP MENUS

Emerson 54e — LOCATING THE CONTROLLER diagram (page 5)
Fig. 5

All menus shown. Some menus will change or not appear with different settings.

Page 6 of the Emerson 54e user manual — STEP 4 - SINGLE-POINT STANDARDIZATION
Page 6 · click to enlarge
Page 6

MODEL 54e pH/ORP QUICK START

MODEL 54e pH/ORP QUICK START

Use the ↓or key to move the cursor down or up and the → or ←to move the cursor right or left.

STEP 3 - AUTOCAL

Press ENTER (F4) to move to the next menu, SAVE (F4) to store a setting, EDIT (F4) to change a setting, and EXIT (F1) to exit without saving changes.

Press any button. The main menu will appear with CAL- IBRATE highlighted. Press ENTER.

STEP 1 - MEASUREMENT TYPE

BUFFER CALIBRATION will be highlighted. Press ENTER.

Place the sensor in the first buffer and press CONT. Wait until the screen reads "Buf 1 done."

Press any button. he main menu will appear with CALI- BRATE highlighted.

Use the arrow buttons to select the correct buffer. Press CONT.

Use the arrow buttons to select Program. Press ENTER.

Rinse and dry the sensor.

Use the arrow buttons to select Configure. Press ENTER.

Place the sensor in the second buffer and press CONT. Wait until the screen reads "Buf 2 done."

DISPLAY will be highlighted. Press ENTER.

Use the arrow buttons to select the correct buffer. Press CONT.

MEASURE will be highlighted. Press ENTER.

Press EXIT until the main display appears.

Use the arrow buttons to select REDOX or ORP. Press SAVE.

Press EXIT until the main display appears.

STEP 4 - SINGLE-POINT STANDARDIZATION

STEP 2 - OUTPUT SETPOINTS

After the sensor is installed, take a grab sample as close to the sensor as possible.

Press any button. The main menu will appear with CAL- IBRATE highlighted.

Use a calibrated pH instrument with automatic tempera- ture compensation to determine the pH of the grab sam- ple.

Use the arrow buttons to select Program. Press ENTER.

Press any button. The main menu will appear with CAL- IBRATE highlighted. Press ENTER.

Use the arrow buttons to select OUTPUT SETPOINTS. Press ENTER.

Use the arrow buttons to select STANDARDIZE. Press ENTER.

OUTPUT 1 SETPOINTS will be highlighted. Press ENTER. Press CONT.

Press EDIT. Use the arrow buttons to enter the pH value of the sample. Press SAVE.

4 mA will be highlighted. Press EDIT. Use the arrow but- tons to enter the desired value. Press SAVE.

Press EXIT until the main display appears.

Use the arrow buttons to select 20 MA. Press EDIT. Use the arrow buttons to enter the desired value. Press SAVE. Press EXIT.

Use the arrow buttons to select OUTPUT 2 SET- POINTS. Press ENTER. Press CONT.

Enter the 4 mA and 20 mA values as above.

Press EXIT until the main display appears.

Page 7 of the Emerson 54e user manual — MODEL 54e pH/ORP QUICK START
Page 7 · click to enlarge
Page 7

MODEL 54e pH/ORP WIRING

Emerson 54e — MODEL 54e pH/ORP QUICK START diagram (page 7)
Fig. 6 · MODEL 54e pH/ORP QUICK START

Sensor Wiring Diagram

Page 8 of the Emerson 54e user manual — MODEL 54e pH/ORP QUICK START
Page 8 · click to enlarge
Page 8
Emerson 54e — MODEL 54e pH/ORP QUICK START diagram (page 8)
Fig. 7 · MODEL 54e pH/ORP QUICK START
Emerson 54e — MODEL 54e pH/ORP QUICK START diagram (page 8)
Fig. 8 · MODEL 54e pH/ORP QUICK START

Power Input and Relay Output Wiring for Model 54e pH/ORP

Emerson 54e — MODEL 54e pH/ORP QUICK START diagram (page 8)
Fig. 9 · MODEL 54e pH/ORP QUICK START

The right people, the right answers, right now. ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE

http://www.raihome.com

Specifications subject to change without notice.

Credit Cards for U.S. Purchases Only.

2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250

http://www.raihome.com

© Rosemount Analytical Inc. 2010

Emerson Process Management

Sources

References & links

Pages referenced in this manual — manufacturer, support and downloads.

Ask about the Emerson 54e

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the Emerson 54e

Be the first to ask — other owners or our team will answer.

Original document

Emerson 54e manual PDF

8 pages·506 KB·Marine Instruments·PDF
Click any page above to open it here
Page 1 / …
100%
Loading pages…