Emerson Liquid Transmitter (1066)

20 pages·6.3 MB·Marine Instruments·PDF·ENDEFRITNLJA
Emerson Liquid Transmitter (1066)
Table of Contents 19 sections 1066 Liquid Analytical Transmitter · ESSENTIAL INSTRUCTIONS Read this page before proceeding! · WARNING: EXPLOSION HAZARD · NOTICE · … ShowHide
Page 1 of the Emerson Liquid Transmitter (1066) user manual — NOTICE
Page 1 · click to enlarge
Page 1

1066 Liquid Analytical Transmitter

ESSENTIAL INSTRUCTIONS Read this page before proceeding!

Emerson Liquid Transmitter (1066) — 1066 Liquid Analytical Transmitter diagram (page 1)
Fig. 1 · 1066 Liquid Analytical Transmitter

Emerson designs, manufactures, and tests its Rosemount Analytical products to meet many national and international standards. Because these instruments are sophisticated technical products, you must properly install, use, and main- tain them to ensure they continue to operate within their normal specifications. The following instructions must be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount Analytical products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; per- sonal injury; property damage; damage to this instrument; and warranty invalidation.

  • Read all instructions prior to installing, operating, and servicing the product. If this Instruction Manual is not the correct manual, telephone 1-800-854-8257 and the requested manual will be provided. Save this Instruction Manual for future reference.
  • If you do not understand any of the instructions, contact your Emerson representative for clarification.
  • Follow all warnings, cautions, and instructions marked on and supplied with the product.
  • Inform and educate your personnel in the proper installation, operation, and maintenance of the product.
  • Install your equipment as specified in the Installation Instructions of the appropriate Instruction Manual and per applicable local and national codes. Connect all products to the proper electrical and pressure sources.
  • To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the product.
  • When replacement parts are required, ensure that qualified people use replacement parts specified by Rosemount. Unauthorized parts and procedures can affect the product’s performance and place the safe operation of your process at risk. Look alike substitutions may result in fire, electrical hazards, or improper operation.
  • Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being performed by qualified persons, to prevent electrical shock and personal injury.

WARNING: EXPLOSION HAZARD

DO NOT OPEN WHILE CIRCUIT IS LIVE. ONLY CLEAN WITH DAMP CLOTH.

NOTICE

If a Model 475 Universal HART® Communicator is used with these transmitters, the software within the Model 475 may require modification. If a software modification is required, please contact your local Emerson Process Management Service Group or National Response Center at 1-800-654-7768.

Page 2 of the Emerson Liquid Transmitter (1066) user manual — QUICK START GUIDE – 1066 Liquid Analytical Transmitter
Page 2 · click to enlarge
Page 2

QUICK START GUIDE – 1066 Liquid Analytical Transmitter

1. For mechanical installation instructions, see page 8 for panel mounting and page 9 for pipe or wall mounting. 2. Wire the sensor to the main circuit board. See page 10 for wiring instructions. Refer to the sensor instruction sheet for additional details. Make loop power connections. 3. Once connections are secured and verified, apply DC loop power to the transmitter. 4. When the transmitter is powered up for the first time, Quick Start screens appear. Quick Start operating tips are as follows:

a. A highlighted field shows the position of the cursor. b. To move the cursor left or right, use the keys to the left or right of the ENTER key. To scroll up or down or to

increase or decrease the value of a digit use the keys above and below the ENTER key. Use the left or right keys to move the decimal point. c. Press ENTER to store a setting. Press EXIT to leave without storing changes. Pressing EXIT during Quick Start

returns the display to the initial start-up screen (select language). 5. Choose the desired language and press ENTER. 6. Choose measurement and press ENTER.

a. For pH or ORP, choose preamplifier location. Select Analyzer to use the integral preamplifier in the transmitter;

select Sensor/J-Box if your sensor is SMART or has an integral preamplifier or if you are using a remote pream- plifier located in a junction box. 5. If applicable, choose units of measurement. 6. For contacting and toroidal conductivity, choose the sensors type and enter the numeric cell constant using the keys. 7. Choose temperature units: °C or °F. 8. After the last step, the main display appears. The outputs are assigned to default values. 9. To change output settings, to scale the 4-20mA current outputs, to change measurement-related settings from the default values, and to enable pH diagnostics, press MENU. Select Program and follow the prompts. Refer to the appropriate menu. 10. To return the transmitter to the factory default settings, choose Program under the main menu, and then scroll to Reset. 11. Please call the Rosemount Analytical Customer Support Center at 1-800-854-8257 if you need further support.

Page 3 of the Emerson Liquid Transmitter (1066) user manual — ATEX
Page 3 · click to enlarge
Page 3

Specifications

GENERAL SPECIFICATIONS CasePolycarbonate. IP66 (CSA, FM), NEMA 4X (CSA)

Dimensions: Overall 155 x 155 x 131mm (6.10 x 6.10 x 5.15 in.). Cutout: 1/2 DIN 139mm x 139mm (5.45 x 5.45 in.)

Conduit openingsSix. Accepts PG13.5 or 1/2 in. conduit fittings

Display: Monochromatic graphic liquid crystal display. No backlight. 128 x 96 pixel display resolution. Active display area: 58 x 78mm (2.3 x 3.0 in.). All fields of the main instrument display can be customized to meet user requirements.

Ambient temperature and humidity: -20 to 65°C (-4 to 149°F), RH 5 to 95% (non-condensing).

Storage Temperature-20 to 70°C (-4 to 158°F)

HART® Communications: PV, SV, TV, and 4V assignable to measurement, temperature and all live HART diagnostics.

RFI/EMI: EN-61326

Complies with the following Standards:

CSA: C22.2 No 0 – 10; C22.2 No 0.4 – 04; C22.2 No. 25-M1966: , C22.2 No. 94-M91: , C22.2 No.142-M1987: , C22.2 No. 157-M1992: , C22.2 No. 213-M1987: , C22.2 No. 60529:05. UL: 50; 508; 913; 1203. ANSI/ISA: 12.12.02-2011.

ATEX: IEC 60079-0:2011, 60079-11:2011

IECExIEC 60079-0: 2011 Edition: 6.0, I EC 60079-11 : 2011-06 Edition: 6.0

FM: 3600: 2011, 3610: 2010, 3611: 2004, 3810: 2005, IEC 60529:2004, ANSI/ISA 60079-0: 2009, ANSI/ISA 60079-11: 2009

Intrinsic Safety (with appropriate safety barrier):

Class I, II, III, Div. 1 Groups A-G T4 Tamb = -20°C to 65°C

IECEx BAS 11.90098X EEx ia IIC T4 Tamb = -20°C to 65°C

Class I, II & III, Division 1, Groups A-G T4 Tamb = -20°C to 40°C for -FI option Tamb = -20°C to 65°C for -HT and -FF options Class I, Zone 0, AEx ia IIC T4 Tamb = -20°C to 40°C for -FI option Tamb = -20°C to 65°C for -HT and -FF options

ATEX

1180 II 1 G Baseefa04ATEX0195X EEx ia IIC T4 Tamb = -20°C to 65°C

Non-Incendive:

Class I, Div. 2, Groups A-D Dust Ignition Proof Class II & III, Div. 1, Groups E-G NEMA 4/4X, IP66 Enclosure T4 Tamb = -20°C to 65°C

Class I, Division 2 Groups A-D Dust Ignition proof Class II & III, Division 1, Groups E-G IP66 enclosure Tamb = -20°C to 40°C for -FI option Tamb = -20°C to 65°C for -HT and –FF options

Hazardous Location Approvals

Page 4 of the Emerson Liquid Transmitter (1066) user manual — Specifications
Page 4 · click to enlarge
Page 4

Input: One isolated sensor input. Measurement choices of pH/ORP, resistivity/conductivity/TDS, % concentration, total and free chlorine, monochloramine, dissolved oxy- gen, dissolved ozone, and temperature. For contacting conductivity measurements, temperature element can be a PT1000 RTD or a PT100 RTD. Other measurements (except ORP) and use PT100 or PT1000 RTDs or a 22k NTC (D.O. only).

Vdc (30 Vdc for intrinsically safe operation). The graph shows the supply voltage required to maintain 12 Vdc (upper line) and 30 Vdc (lower line) at the transmitter ter- minals when the current is 22 mA.

Analog Outputs: Two-wire loop powered (Output 1 only). Two 4-20 mA electrically isolated current outputs (Output 2 must be externally powered). Superimposed HART digital signal on Output 1. Fully scalable over the operating range of the sensor.

Power & Load Requirements: Supply voltage at the transmitter terminals should be at least 12.7Vdc. Power supply voltage should cover the voltage drop on the cable plus the external load resistor required for HART commu- nications (250 Ω minimum). Minimum power supply volt- age is 12.7Vdc. Maximum power supply voltage is 42.4

Weight/Shipping Weight2 lbs/3 lbs (1 kg/1.5 kg)
Emerson Liquid Transmitter (1066) — Specifications diagram (page 4)
Fig. 2 · Specifications

12 18 24 30 36 42

Power supply voltage, Vdc

FIGURE 2. Power Supply-Current Loop Wiring FIGURE 1. Load/Power Supply Requirements

Show 6 more labels / short text on this page
1364 ohms · Load, ohms · with HART communication · 545 ohms · without HART communication · HART option
Page 5 of the Emerson Liquid Transmitter (1066) user manual — PERFORMANCE SPECIFICATIONS Recommended Range – Contacting Conductivity
Page 5 · click to enlarge
Page 5

Specifications CONTINUED

CONTACTING CONDUCTIVITY Performance Specifications

Measurement Rangesee table below

Input filter: time constant 1 - 999 sec, default 2 sec.

Response time: 3 seconds to 95% of final reading using the default input filter Recommended Sensors for Conductivity All Rosemount Analytical ENDURANCE Model 400 series conductivity sensors (Pt 1000 RTD) and Model 410VP 4- electrode sensor.

PERFORMANCE SPECIFICATIONS Recommended Range – Contacting Conductivity

Cell 0.01S/cm 0.1mS/cm 1.0mS/cm 10mS/cm 100mS/cm 1000mS/cm 10mS/cm 100mS/cm 1000mS/cm Constant 0.010.01mS/cm to 200mS/cm 200mS/cm to 2000mS/cm 0.10.1mS/cm to 2000mS/cm 2000mS/cm to 20mS/cm 1.01 mS/cm to 20mS/cm 20mS/cm to 200mS/cm 4-electrode2mS/cm to 2mS/cm to1400mS/cm

Linearity for Standard Cable ≤50 ft (15 m)

±0.6% of reading in recommended range

±2% of reading outside high recommended range

±5% of reading outside low recommended range

±4% of reading in recommended range

Temperature specifications:

Temperature range0 to 200°C Temperature Accuracy, Pt-1000, 0-50°C± 0.1°C Temperature Accuracy, Pt-1000, Temp. > 50°C± 0.5°C
Page 6 of the Emerson Liquid Transmitter (1066) user manual — PERFORMANCE SPECIFICATIONS Recommended Range - Toroidal Conductivity
Page 6 · click to enlarge
Page 6

Specifications CONTINUED

TOROIDAL CONDUCTIVITY Performance Specifications

Measurement Rangesee table below

Input filter: time constant 1 - 999 sec, default 2 sec.

Response time: 3 seconds to 95% of final reading Recommended Sensors for Conductivity All Rosemount Analytical submersion/immersion and flow-through toroidal sensors.

PERFORMANCE SPECIFICATIONS Recommended Range - Toroidal Conductivity

Model 1mS/cm 10mS/cm 100mS/cm 1000mS/cm 10mS/cm 100mS/cm 1000mS/cm 2000mS/cm

5mS/cm to 500mS/cm

15mS/cm to 1500mS/cm

100mS/cm to 2000mS/cm

222 (1in & 2in)

500mS/cm to 2000mS/cm

LOOP PERFORMANCE (Following Calibration)

Model 226: ±1% of reading ±5mS/cm in recommended range Models 225 & 228: ±1% of reading ±15mS/cm in recommended range Models 222, 242: ±4% of reading ±5mS/cm in recommended range

Models 225, 226 & 228±5% of reading outside high recommended range

Temperature specifications:

Temperature range-25 to 210°C (-13 to 410ºF) Temperature Accuracy, Pt-100, -25 to 50 °C± 0.5°C Temperature Accuracy, Pt-100,. 50 to 210°C± 1°C

500mS/cm to 2000mS/cm · 225 & 228 · 1500mS/cm to 2000mS/cm

Page 7 of the Emerson Liquid Transmitter (1066) user manual — WARNING: RISK OF ELECTRICAL SHOCK
Page 7 · click to enlarge
Page 7

Installation

UNPACKING AND INSPECTION Inspect the shipping container. If it is damaged, contact the shipper immediately for instructions. Save the box. If there is no apparent damage, unpack the container. Be sure all items shown on the packing list are present. If items are miss- ing, notify Rosemount Analytical immediately.

INSTALLATION General Information 1. Although the transmitter is suitable for outdoor use, installation is direct sunlight or in areas of extreme tempera- tures is not recommended unless a sunshield is used.

2. Install the transmitter in an area where vibration and electromagnetic and radio frequency interference are minimized or absent.
3. Keep the transmitter and sensor wiring at least one foot from high voltage conductors. Be sure there is easy access to the transmitter.
4. The transmitter is suitable for panel, pipe, or surface mounting.
5. The transmitter case has six 1/2-inch (PG13.5) conduit openings. Use separate conduit openings for the power/output cable, the sensor cable, and the other the sensor cable as needed (pH input for free chlorine with continuous pH correction).
6. Use weathertight cable glands to keep moisture out to the transmitter. If conduit is used, plug and seal the connections at the transmitter housing to prevent moisture from getting inside the instrument.

PREPARING CONDUIT OPENINGS There are six conduit openings in all configurations of Model 1066. (Note: four enclosure opening plugs will be provid- ed upon shipment.) Conduit openings accept 1/2-inch conduit fittings or PG13.5 cable glands. To keep the case watertight, block unused openings with NEMA 4X or IP66 conduit plugs. NOTE: Use watertight fittings and hubs that comply with your requirements. Connect the conduit hub to the conduit before attaching the fitting to the transmitter.

WARNING: RISK OF ELECTRICAL SHOCK

Electrical installation must be in accordance with the National Electrical Code (ANSI/NFPA-70) and/or any other applicable

national or local codes.

ELECTROSTATIC IGNITION HAZARD Special condition for safe use (when installed in hazardous area) 1. The plastic enclosure, excepting the front panel, must only be cleaned with a damp cloth. The surface resistivity of the non-metallic enclosure materials is greater than one gigaohm. Care must be taken to avoid electrostatic charge build-up. The Model 1066 Transmitter must not be rubbed or cleaned with solvents or a dry cloth.

2. The panel mount gasket has not been tested for type of protection IP66 or Class II and III. Type of protection IP66 and Class II, III refer the enclosure only.
3. The surface resistivity of the non-metallic enclosure materials is greater than one gigaohm. Care must be taken to avoid electrostatic charge build-up. The Model 1066 Transmitter must not be rubbed or cleaned with solvents or a dry cloth.
4. Special Condition of Use of 1066-C-FF/FI-67 and 1066-T-FF/FI-67. For use with simple apparatus model series 140, 141, 142, 150, 400, 401, 402, 402VP, 403, 403VP, 404, and 410VP contacting conductivity sensors and model series 222, 225, 226, 228 toroidal sensors.
Page 8 of the Emerson Liquid Transmitter (1066) user manual — Installation
Page 8 · click to enlarge
Page 8

FIGURE 3. Panel Mounting Dimensions

Emerson Liquid Transmitter (1066) — Installation diagram (page 8)
Fig. 3 · Installation
Page 9 of the Emerson Liquid Transmitter (1066) user manual — Installation
Page 9 · click to enlarge
Page 9

FIGURE 4. Pipe and wall mounting dimensions (Mounting bracket PN: 23820-00)

Emerson Liquid Transmitter (1066) — Installation diagram (page 9)
Fig. 4 · Installation
Page 10 of the Emerson Liquid Transmitter (1066) user manual — Installation
Page 10 · click to enlarge
Page 10

FIGURE 5. Contacting and Toroidal Conductivity sensor wiring to 1066 circuit board (1066-C and 1066-T)

Emerson Liquid Transmitter (1066) — Installation diagram (page 10)
Fig. 5 · Installation
Page 11 of the Emerson Liquid Transmitter (1066) user manual — Installation
Page 11 · click to enlarge
Page 11

FIGURE 6. CSA Installation

Emerson Liquid Transmitter (1066) — Installation diagram (page 11)
Fig. 6 · Installation
Page 12 of the Emerson Liquid Transmitter (1066) user manual — Installation
Page 12 · click to enlarge
Page 12

FIGURE 7. CSA Installation

Emerson Liquid Transmitter (1066) — Installation diagram (page 12)
Fig. 7 · Installation
Page 13 of the Emerson Liquid Transmitter (1066) user manual — Installation
Page 13 · click to enlarge
Page 13

FIGURE 8. CSA Installation

Emerson Liquid Transmitter (1066) — Installation diagram (page 13)
Fig. 8 · Installation
Page 14 of the Emerson Liquid Transmitter (1066) user manual — LABEL, INFO, 1066
Page 14 · click to enlarge
Page 14

FIGURE 9. CSA Label Information

Emerson Liquid Transmitter (1066) — LABEL, INFO, 1066 diagram (page 14)
Fig. 9

LABEL, INFO, 1066

CSA

Emerson Liquid Transmitter (1066) — LABEL, INFO, 1066 diagram (page 14)
Fig. 10
Page 15 of the Emerson Liquid Transmitter (1066) user manual — LABEL, INFO, 1066
Page 15 · click to enlarge
Page 15

FIGURE 10. ATEX, IECEx Label Information

Emerson Liquid Transmitter (1066) — LABEL, INFO, 1066 diagram (page 15)
Fig. 11

LABEL, INFO, 1066

ATEX

9241717-00

Page 16 of the Emerson Liquid Transmitter (1066) user manual — LABEL, INFO, 1066
Page 16 · click to enlarge
Page 16

FIGURE 11. FM installation

Emerson Liquid Transmitter (1066) — LABEL, INFO, 1066 diagram (page 16)
Fig. 12
Page 17 of the Emerson Liquid Transmitter (1066) user manual — LABEL, INFO, 1066
Page 17 · click to enlarge
Page 17

FIGURE 12. FM Installation

Emerson Liquid Transmitter (1066) — LABEL, INFO, 1066 diagram (page 17)
Fig. 13
Page 18 of the Emerson Liquid Transmitter (1066) user manual — LABEL, INFO, 1066
Page 18 · click to enlarge
Page 18

FIGURE 13. FM installation

Emerson Liquid Transmitter (1066) — LABEL, INFO, 1066 diagram (page 18)
Fig. 14
Page 19 of the Emerson Liquid Transmitter (1066) user manual — LABEL, INFO, 1066
Page 19 · click to enlarge
Page 19

FIGURE 14. FM label information

Emerson Liquid Transmitter (1066) — FM diagram (page 19)
Fig. 15

LABEL, INFO, 1066

FM

Emerson Liquid Transmitter (1066) — LABEL, INFO, 1066 diagram (page 19)
Fig. 16
Page 20 of the Emerson Liquid Transmitter (1066) user manual — FM
Page 20 · click to enlarge
Page 20
Emerson Liquid Transmitter (1066) — FM diagram (page 20)
Fig. 17
Emerson Liquid Transmitter (1066) — FM diagram (page 20)
Fig. 18

facebook.com/EmersonRosemountAnalytical

AnalyticExpert.com

Credit Cards for U.S. Purchases Only.

twitter.com/RAIhome

youtube.com/user/RosemountAnalytical

©2013 Rosemount Analytical, Inc. All rights reserved.

2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250

The Emerson logo is a trademark and service mark of Emerson Electric Co. Brand name is a mark of one of the Emerson Process Management family of companies. All other marks are the property of their respective owners.

The contents of this publication are presented for information purposes only, and while effort has been made to ensure their accuracy, they are not to be construed as warranties or guarantees, express or implied, regarding the products or services described herein or their use or applicability. All sales are governed by our terms and conditions, which are available on request. We reserve the right to modify or improve the designs or specifications of our products at any time without notice.

rosemountanalytical.com

© Rosemount Analytical Inc. 2013 Rev. A

Emerson Process Management

Sources

References & links

Pages referenced in this manual — manufacturer, support and downloads.

Ask about the Emerson Liquid Transmitter (1066)

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the Emerson Liquid Transmitter (1066)

Be the first to ask — other owners or our team will answer.

Original document

Emerson Liquid Transmitter (1066) manual PDF

20 pages·6.3 MB·Marine Instruments·PDF
Click any page above to open it here