ESSENTIAL INSTRUCTIONS Read this page before proceeding!

Emerson designs, manufactures, and tests its Rosemount Analytical products to meet many national and international standards. Because these instruments are sophisticated technical products, you must properly install, use, and main- tain them to ensure they continue to operate within their normal specifications. The following instructions must be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount Analytical products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; per- sonal injury; property damage; damage to this instrument; and warranty invalidation.
DO NOT OPEN WHILE CIRCUIT IS LIVE. ONLY CLEAN WITH DAMP CLOTH.
If a Model 475 Universal HART® Communicator is used with these transmitters, the software within the Model 475 may require modification. If a software modification is required, please contact your local Emerson Process Management Service Group or National Response Center at 1-800-654-7768.

1. For mechanical installation instructions, see page 8 for panel mounting and page 9 for pipe or wall mounting. 2. Wire the sensor to the main circuit board. See page 10 for wiring instructions. Refer to the sensor instruction sheet for additional details. Make loop power connections. 3. Once connections are secured and verified, apply DC loop power to the transmitter. 4. When the transmitter is powered up for the first time, Quick Start screens appear. Quick Start operating tips are as follows:
a. A highlighted field shows the position of the cursor. b. To move the cursor left or right, use the keys to the left or right of the ENTER key. To scroll up or down or to
increase or decrease the value of a digit use the keys above and below the ENTER key. Use the left or right keys to move the decimal point. c. Press ENTER to store a setting. Press EXIT to leave without storing changes. Pressing EXIT during Quick Start
returns the display to the initial start-up screen (select language). 5. Choose the desired language and press ENTER. 6. Choose measurement and press ENTER.
a. For pH or ORP, choose preamplifier location. Select Analyzer to use the integral preamplifier in the transmitter;
select Sensor/J-Box if your sensor is SMART or has an integral preamplifier or if you are using a remote pream- plifier located in a junction box. 5. If applicable, choose units of measurement. 6. For contacting and toroidal conductivity, choose the sensors type and enter the numeric cell constant using the keys. 7. Choose temperature units: °C or °F. 8. After the last step, the main display appears. The outputs are assigned to default values. 9. To change output settings, to scale the 4-20mA current outputs, to change measurement-related settings from the default values, and to enable pH diagnostics, press MENU. Select Program and follow the prompts. Refer to the appropriate menu. 10. To return the transmitter to the factory default settings, choose Program under the main menu, and then scroll to Reset. 11. Please call the Rosemount Analytical Customer Support Center at 1-800-854-8257 if you need further support.

Dimensions: Overall 155 x 155 x 131mm (6.10 x 6.10 x 5.15 in.). Cutout: 1/2 DIN 139mm x 139mm (5.45 x 5.45 in.)
Display: Monochromatic graphic liquid crystal display. No backlight. 128 x 96 pixel display resolution. Active display area: 58 x 78mm (2.3 x 3.0 in.). All fields of the main instrument display can be customized to meet user requirements.
Ambient temperature and humidity: -20 to 65°C (-4 to 149°F), RH 5 to 95% (non-condensing).
HART® Communications: PV, SV, TV, and 4V assignable to measurement, temperature and all live HART diagnostics.
Complies with the following Standards:
CSA: C22.2 No 0 – 10; C22.2 No 0.4 – 04; C22.2 No. 25-M1966: , C22.2 No. 94-M91: , C22.2 No.142-M1987: , C22.2 No. 157-M1992: , C22.2 No. 213-M1987: , C22.2 No. 60529:05. UL: 50; 508; 913; 1203. ANSI/ISA: 12.12.02-2011.
FM: 3600: 2011, 3610: 2010, 3611: 2004, 3810: 2005, IEC 60529:2004, ANSI/ISA 60079-0: 2009, ANSI/ISA 60079-11: 2009
Intrinsic Safety (with appropriate safety barrier):
Class I, II, III, Div. 1 Groups A-G T4 Tamb = -20°C to 65°C
IECEx BAS 11.90098X EEx ia IIC T4 Tamb = -20°C to 65°C
Class I, II & III, Division 1, Groups A-G T4 Tamb = -20°C to 40°C for -FI option Tamb = -20°C to 65°C for -HT and -FF options Class I, Zone 0, AEx ia IIC T4 Tamb = -20°C to 40°C for -FI option Tamb = -20°C to 65°C for -HT and -FF options
1180 II 1 G Baseefa04ATEX0195X EEx ia IIC T4 Tamb = -20°C to 65°C
Non-Incendive:
Class I, Div. 2, Groups A-D Dust Ignition Proof Class II & III, Div. 1, Groups E-G NEMA 4/4X, IP66 Enclosure T4 Tamb = -20°C to 65°C
Class I, Division 2 Groups A-D Dust Ignition proof Class II & III, Division 1, Groups E-G IP66 enclosure Tamb = -20°C to 40°C for -FI option Tamb = -20°C to 65°C for -HT and –FF options
Hazardous Location Approvals

Input: One isolated sensor input. Measurement choices of pH/ORP, resistivity/conductivity/TDS, % concentration, total and free chlorine, monochloramine, dissolved oxy- gen, dissolved ozone, and temperature. For contacting conductivity measurements, temperature element can be a PT1000 RTD or a PT100 RTD. Other measurements (except ORP) and use PT100 or PT1000 RTDs or a 22k NTC (D.O. only).
Vdc (30 Vdc for intrinsically safe operation). The graph shows the supply voltage required to maintain 12 Vdc (upper line) and 30 Vdc (lower line) at the transmitter ter- minals when the current is 22 mA.
Analog Outputs: Two-wire loop powered (Output 1 only). Two 4-20 mA electrically isolated current outputs (Output 2 must be externally powered). Superimposed HART digital signal on Output 1. Fully scalable over the operating range of the sensor.
Power & Load Requirements: Supply voltage at the transmitter terminals should be at least 12.7Vdc. Power supply voltage should cover the voltage drop on the cable plus the external load resistor required for HART commu- nications (250 Ω minimum). Minimum power supply volt- age is 12.7Vdc. Maximum power supply voltage is 42.4

12 18 24 30 36 42
Power supply voltage, Vdc
FIGURE 2. Power Supply-Current Loop Wiring FIGURE 1. Load/Power Supply Requirements

Input filter: time constant 1 - 999 sec, default 2 sec.
Response time: 3 seconds to 95% of final reading using the default input filter Recommended Sensors for Conductivity All Rosemount Analytical ENDURANCE Model 400 series conductivity sensors (Pt 1000 RTD) and Model 410VP 4- electrode sensor.
Linearity for Standard Cable ≤50 ft (15 m)
±0.6% of reading in recommended range
±2% of reading outside high recommended range
±5% of reading outside low recommended range
±4% of reading in recommended range
Temperature specifications:

Input filter: time constant 1 - 999 sec, default 2 sec.
Response time: 3 seconds to 95% of final reading Recommended Sensors for Conductivity All Rosemount Analytical submersion/immersion and flow-through toroidal sensors.
Model 1mS/cm 10mS/cm 100mS/cm 1000mS/cm 10mS/cm 100mS/cm 1000mS/cm 2000mS/cm
5mS/cm to 500mS/cm
15mS/cm to 1500mS/cm
100mS/cm to 2000mS/cm
222 (1in & 2in)
500mS/cm to 2000mS/cm
LOOP PERFORMANCE (Following Calibration)
Model 226: ±1% of reading ±5mS/cm in recommended range Models 225 & 228: ±1% of reading ±15mS/cm in recommended range Models 222, 242: ±4% of reading ±5mS/cm in recommended range
Temperature specifications:
500mS/cm to 2000mS/cm · 225 & 228 · 1500mS/cm to 2000mS/cm

UNPACKING AND INSPECTION Inspect the shipping container. If it is damaged, contact the shipper immediately for instructions. Save the box. If there is no apparent damage, unpack the container. Be sure all items shown on the packing list are present. If items are miss- ing, notify Rosemount Analytical immediately.
INSTALLATION General Information 1. Although the transmitter is suitable for outdoor use, installation is direct sunlight or in areas of extreme tempera- tures is not recommended unless a sunshield is used.
PREPARING CONDUIT OPENINGS There are six conduit openings in all configurations of Model 1066. (Note: four enclosure opening plugs will be provid- ed upon shipment.) Conduit openings accept 1/2-inch conduit fittings or PG13.5 cable glands. To keep the case watertight, block unused openings with NEMA 4X or IP66 conduit plugs. NOTE: Use watertight fittings and hubs that comply with your requirements. Connect the conduit hub to the conduit before attaching the fitting to the transmitter.
Electrical installation must be in accordance with the National Electrical Code (ANSI/NFPA-70) and/or any other applicable
national or local codes.
ELECTROSTATIC IGNITION HAZARD Special condition for safe use (when installed in hazardous area) 1. The plastic enclosure, excepting the front panel, must only be cleaned with a damp cloth. The surface resistivity of the non-metallic enclosure materials is greater than one gigaohm. Care must be taken to avoid electrostatic charge build-up. The Model 1066 Transmitter must not be rubbed or cleaned with solvents or a dry cloth.

FIGURE 3. Panel Mounting Dimensions


FIGURE 4. Pipe and wall mounting dimensions (Mounting bracket PN: 23820-00)


FIGURE 5. Contacting and Toroidal Conductivity sensor wiring to 1066 circuit board (1066-C and 1066-T)


FIGURE 6. CSA Installation


FIGURE 7. CSA Installation


FIGURE 8. CSA Installation


FIGURE 9. CSA Label Information

CSA


FIGURE 10. ATEX, IECEx Label Information

ATEX
9241717-00

FIGURE 11. FM installation


FIGURE 12. FM Installation


FIGURE 13. FM installation


FIGURE 14. FM label information

FM




facebook.com/EmersonRosemountAnalytical
Credit Cards for U.S. Purchases Only.
youtube.com/user/RosemountAnalytical
©2013 Rosemount Analytical, Inc. All rights reserved.
2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250
The Emerson logo is a trademark and service mark of Emerson Electric Co. Brand name is a mark of one of the Emerson Process Management family of companies. All other marks are the property of their respective owners.
The contents of this publication are presented for information purposes only, and while effort has been made to ensure their accuracy, they are not to be construed as warranties or guarantees, express or implied, regarding the products or services described herein or their use or applicability. All sales are governed by our terms and conditions, which are available on request. We reserve the right to modify or improve the designs or specifications of our products at any time without notice.
© Rosemount Analytical Inc. 2013 Rev. A
Emerson Process Management
Pages referenced in this manual — manufacturer, support and downloads.
Owners and our team usually reply within a day — and your question helps the next person too.
Be the first to ask — other owners or our team will answer.