Model 404
ENDURANCE
™ Conductivity Sensor
For additional information, please visit our website
at www.emersonprocess.com/raihome/liquid/.
WARNING
WARNING
Before removing the sensor, be absolutely certain that the process pressure is reduced to 0 psig and the process temperature is lowered to a safe level!
The wetted sensor materials may not be compatible with process com position and operating conditions. Application compat ibility is entirely the responsi- bility of the user.



FIGURE 2. Model 54eC Wiring FIGURE 3. Model 56 and 1056 Wiring



FIGURE 5. Model Xmt-C-10 Wiring (Panel)
FIGURE 6. Model Xmt-C-11 Wiring (Pipe or Wall)


FIGURE 7. Model 5081-C Wiring
FIGURE 8. Model 6081-C Wiring
If wiring connections are made through a remote junction box (PN 23550-00), wire point-to-point. Use cable 23747-00 (factory-terminated) or 9200275 (no terminations).

CLEANING THE SENSOR
The 404-17 (stainless steel body) sensor can be taken apart for cleaning. However, in some cases, disassembling and reassembling the sensor can cause the cell constant to change as much as 1%. For maximum accuracy, the cell con- stant should be rechecked after the sensor has been reassembled. The 400-16 (PVC body) sensor cannot be taken apart.
Use a warm detergent solution and a soft brush or pipe cleaner to remove oil and scale. Isopropyl alcohol (rubbing alco- hol) can also be used to remove oily films. Avoid using strong mineral acids to clean conductivity sensors.
ENDURANCE conductivity sensors are calibrated at the factory and do not need calibration when first placed in service. Simply enter the cell constant printed on the label into the analyzer.
After a period of service, the sensor may require calibration. Because Model 404 sensors have a flow-through design, they are best calibrated against a referee meter and sensor where the two sensors are connected in series with the same liquid flowing through both.
For more information about calibrating contacting conductivity sensors, refer to application sheet ADS 43-024, available on the Rosemount Analytical website.

TROUBLESHOOTING
FIGURE 9. Checking Temperature Element Disconnect leads and measure resistance shown. The meas- ured resistance should be close to the value in the table.
FIGURE 10. Checking Continuity and Leakage Disconnect electrode leads and measure resistance and continuity as shown. Sensor must be dry when checking resistance between electrode leads.




The right people, the right answers, right now. ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE
Specifications subject to change without notice.
Credit Cards for U.S. Purchases Only.
2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250
© Rosemount Analytical Inc. 2010
Emerson Process Management
Pages referenced in this manual — manufacturer, support and downloads.
Common problems, likely causes, and what to do.
| Problem | Possible cause | Remedy | |
|---|---|---|---|
| Off-scale reading | Wiring is wrong. | Verify wiring. | p.5 |
| Off-scale reading | Temperature element is open or shorted. | Check temperature element for open or short circuits. See Figure 9. | p.5 |
| Off-scale reading | Sensor is not in process stream. | Be sure sensor is completely submerged in process stream. | p.5 |
| Off-scale reading | Variopol cable is not properly seated. | Loosen connector and reseat. | p.5 |
| Off-scale reading | Sensor has failed. | Perform isolation checks. See Figure 10. | p.5 |
| Noisy reading | Sensor is improperly installed in process stream. | Be sure sensor is completely submerged in process stream. | p.5 |
| Noisy reading | Variopol cable is not properly seated. | Loosen connector and reseat. | p.5 |
| Reading seems wrong (lower or higher than expected) | Bubbles trapped in sensor. | Be sure sensor is properly oriented in pipe or flow cell. See Figure 1. Apply back pressure to flow cell. | p.5 |
| Reading seems wrong (lower or higher than expected) | Wrong temperature correction algorithm. | Check that temperature correction is appropriate for the sample. See analyzer manual for more information. | p.5 |
| Reading seems wrong (lower or higher than expected) | Wrong cell constant. | Verify that the correct cell constant has been entered in the analyzer and that the cell constant is appropriate for the conductivity of the sample. See analyzer manual. | p.5 |
| Sluggish response | Electrodes are fouled. | Clean electrodes. | p.5 |
| Sluggish response | Sensor is sampling a dead area. | Move sample line to a location more representative of the process liquid. | p.5 |