DanielTM 3812 Liquid Ultrasonic Flow Meter. Installation Manualp.1Contentsp.9Prefacep.3Daniel customer servicep.3Signal words and symbolsp.4Important safety instructionsp.5Noticep.7Warranty and Limitationsp.8List of Tablesp.11List of Figuresp.13Section 1: Introductionp.151.1 Typical Applicationsp.151.2 Features and benefitsp.161.3 Acronyms, abbreviations and definitionsp.171.4 Daniel MeterLink softwarep.191.5 Daniel 3812 Liquid Ultrasonic Flow meter designp.201.6 Meter specificationsp.271.7 Pre-installation considerationsp.311.8 Safetyp.311.9 Daniel 3812 Certifications and Approvalsp.331.10 FCC compliancep.33Section 2: Mechanical installationp.352.1 Meter piping, lifting and mountingp.352.2 Piping recommendationsp.422.3 Meter safety for hoist rings and lifting slingsp.462.3.1 Use of appropriate safety engineered swivel hoist rings in meter end flangesp.472.3.2 Appropriately rated lifting slingsp.542.4 Mounting requirements in heated or cooled pipelinesp.57Section 3: Electrical installationp.593.1 Cable length TTL modep.593.2 Cable length Open Collector modep.593.3 Grounding meter electronics housingp.603.4 Conduit sealsp.613.4.1 Startup for systems using explosion-proof conduitp.623.4.2 Startup for systems that use flame-proof cablep.633.5 Wiring and I/Op.643.5.1 CPU Module labeling and LED indicatorsp.653.6 I/O connectionsp.693.6.1 Frequency/Digital outputsp.703.6.2 Analog input settingsp.743.6.3 Analog output settingsp.743.6.4 Digital Inputp.743.6.5 DHCP server switch settingsp.743.6.6 Configuration protect switch settingsp.753.6.7 External power source connection and fusep.753.7 Security seal installationp.763.7.1 Direct or remote mount transmitter electronics enclosure sealp.763.7.2 Base Enclosure Security Sealsp.783.7.3 Seal the Junction Box remote mount electronics optionp.793.7.4 Bolted band shroud security sealsp.803.7.5 Clamped band shroud security sealsp.813.7.6 Split shroud security sealsp.823.7.7 Latched band shroud security sealsp.833.7.8 Seal conduit portsp.85Section 4: Configurationp.874.1 Daniel MeterLink setupp.874.2 Field Setup Wizardp.884.3 Using AMS Device Manager to configure the meterp.924.4 Using a Field Communicator to configure the meterp.1074.5 Security seals for the meterp.110Appendix A: Engineering drawingsp.111A.1 Daniel 3812 Liquid Ultrasonic Flow Meter drawingsp.111Appendix B: Open source licensesp.117B.1 GNU General Public Licensep.118B.2 GNU Lesser General Public Licensep.129B.3 BSD Open Source Licensep.133B.4 M.I.T Licensep.134Appendix C Indexp.135C.1 Manual Indexp.135
Page 1 of the Emerson 3812 user manual — DanielTM 3812 Liquid Ultrasonic Flow Meter. Installation Manual
Page 1 · click to enlarge
Page 1

Installation Manual Pa r t N u m b er 3 - 9 0 00 - 76 5 Rev i s i o n H

J u n e 2 0 15 DanielTM 3812 Liquid Ultrasonic Flow Meter for Direct or Remote Mount Meter Electronics

Emerson 3812 — DanielTM 3812 Liquid Ultrasonic Flow Meter. Installation Manual diagram (page 1)
Fig. 1 · DanielTM 3812 Liquid Ultrasonic Flow Meter. Installa…
Emerson 3812 — DanielTM 3812 Liquid Ultrasonic Flow Meter. Installation Manual diagram (page 1)
Fig. 2 · DanielTM 3812 Liquid Ultrasonic Flow Meter. Installa…
Page 2 of the Emerson 3812 user manual — DanielTM 3812 Liquid Ultrasonic Flow Meter. Installation Manual
Page 2 · click to enlarge
Page 2
Page 2 of the Emerson 3812 user manual — DanielTM 3812 Liquid Ultrasonic Flow Meter. Installation Manual
Page 3 of the Emerson 3812 user manual — Email
Page 3 · click to enlarge
Page 3

Daniel customer service

LocationTelephone numberFax number North America/Latin America+1.713.467.6000+1.713.827.4805 Daniel Customer Service+1.713.827.6314+1.713.827.6312 USA (toll free)+1.888.356.9001+1.713.827.3380 Asia Pacific (Republic of Singapore)+65.6777.8211+65.6777.0947.0743 Europe (Stirling Scotland, UK)+44 (0)1786.433400+44 (0)1786.433401 Middle East Africa (Dubai, UAE)+971 4 8118100+971 4 8865465

Email

  • Customer Service: tech.service@<a href="http://emersonprocess.com" rel="nofollow noopener" target="_blank">emersonprocess.com</a>
  • Customer Support: daniel.cst.support@<a href="http://emerson.com" rel="nofollow noopener" target="_blank">emerson.com</a>
  • Asia-Pacific: danielap.support@<a href="http://emerson.com" rel="nofollow noopener" target="_blank">emerson.com</a>
  • Europe: DanielEMA.CST@<a href="http://EmersonProcess.com" rel="nofollow noopener" target="_blank">EmersonProcess.com</a>

Return Material Authorization (RMA)

A Return Material Authorization (RMA) number must be obtained prior to returning any equipment for any reason. Download the RMA form from the Support Services web page by selecting the link below.

www2.emersonprocess.com/EN-US/BRANDS/DANIEL/SUPPORT-SERVICES/Pages/Support-Services.aspx?

Page 4 of the Emerson 3812 user manual — Note
Page 4 · click to enlarge
Page 4

Signal words and symbols

Pay special attention to the following signal words, safety alert symbols and statements:

This is a safety alert symbol. It is used to alert you to potential physical injury hazards. Obey all safety messages that follow this symbol to avoid possible injury or death.

Important

Tip

Note

Safety alert symbol

Page 5 of the Emerson 3812 user manual — Important safety instructions
Page 5 · click to enlarge
Page 5

Important safety instructions

Daniel Measurement and Control, Inc. (Daniel) designs, manufactures and tests products to function within specific conditions. Because these products are sophisticated technical instruments, it is important that the owner and operation personnel strictly adhere both to the information printed on the product and to all instructions provided in this manual prior to installation, operation, and maintenance.

Daniel also urges you to integrate this manual into your training and safety program.

BE SURE ALL PERSONNEL READ AND FOLLOW THE INSTRUCTIONS IN THIS MANUAL AND ALL NOTICES AND PRODUCT WARNINGS.

Installing, operating or maintaining a Daniel product improperly could lead to serious injury or death from explosion or exposure to dangerous substances. To reduce this risk:

  • Comply with all information on the product, in this manual, and in any local and national codes that apply to the product.
  • Do not allow untrained personnel to work with this product.
  • Use Daniel parts and work procedures specified in this manual.

Product owners (Purchasers):

  • Use the correct product for the environment and pressures present. See technical data or product specifications for limitations. If you are unsure, discuss your needs with your Daniel representative.
  • Inform and train all personnel in the proper installation, operation, and maintenance of this product.
  • To ensure safe and proper performance, only informed and trained personnel should install, operate, repair and maintain this product.
  • Verify that this is the correct instruction manual for your Daniel product. If this is not the correct documentation, contact Daniel at 1-713-827-6314. You may also download the correct manual from:

http://www.daniel.com

  • Save this instruction manual for future reference.
  • If you resell or transfer this product, it is your responsibility to forward this instruction manual along with the product to the new owner or transferee.
  • ALWAYS READ AND FOLLOW THE INSTALLATION, OPERATIONS, MAINTENANCE AND TROUBLESHOOTING MANUALS AND ALL PRODUCT WARNINGS AND INSTRUCTIONS.
  • Do not use this equipment for any purpose other than its intended service. This may result in property damage and/or serious personal injury or death.
Page 6 of the Emerson 3812 user manual — Important safety instructions
Page 6 · click to enlarge
Page 6

Product Operation Personnel:

  • To prevent personal injury, personnel must follow all instructions of this manual prior to and during operation of the product.
  • Follow all warnings, cautions, and notices marked on, and supplied with, this product.
  • Verify that this is the correct instruction manual for your Daniel product. If this is not the correct documentation, contact Daniel at 1-713-827-6314. You may also download the correct manual from:

http://www.daniel.com

  • Read and understand all instructions and operating procedures for this product.
  • If you do not understand an instruction, or do not feel comfortable following the instructions, contact your Daniel representative for clarification or assistance.
  • Install this product as specified in the INSTALLATION section of this manual per applicable local and national codes.
  • Follow all instructions during the installation, operation, and maintenance of this product.
  • Connect the product to the appropriate pressure and electrical sources when and where applicable.
  • Ensure that all connections to pressure and electrical sources are secure prior to and during equipment operation.
  • Use only replacement parts specified by Daniel. Unauthorized parts and procedures can affect this product's performance, safety, and invalidate the warranty. "Look-a-like" substitutions may result in deadly fire, explosion, release of toxic substances or improper operation.
  • Save this instruction manual for future reference.
Page 7 of the Emerson 3812 user manual — Notice
Page 7 · click to enlarge
Page 7

Notice

THE CONTENTS OF THIS PUBLICATION ARE PRESENTED FOR INFORMATIONAL PURPOSES ONLY, AND WHILE EVERY EFFORT HAS BEEN MADE TO ENSURE THEIR ACCURACY, THEY ARE NOT TO BE CONSTRUED AS WARRANTIES OR GUARANTEES, EXPRESSED OR IMPLIED, REGARDING THE PRODUCTS OR SERVICES DESCRIBED HEREIN OR THEIR USE OR APPLICABILITY. ALL SALES ARE GOVERNED BY DANIEL'S TERMS AND CONDITIONS, WHICH ARE AVAILABLE UPON REQUEST. WE RESERVE THE RIGHT TO MODIFY OR IMPROVE THE DESIGNS OR SPECIFICATIONS OF SUCH PRODUCTS AT ANY TIME.

DANIEL DOES NOT ASSUME RESPONSIBILITY FOR THE SELECTION, USE OR MAINTENANCE OF ANY PRODUCT. RESPONSIBILITY FOR PROPER SELECTION, USE AND MAINTENANCE OF ANY DANIEL PRODUCT REMAINS SOLELY WITH THE PURCHASER AND END-USER.

TO THE BEST OF DANIEL'S KNOWLEDGE THE INFORMATION HEREIN IS COMPLETE AND ACCURATE. DANIEL MAKES NO WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE WITH RESPECT TO THIS MANUAL AND, IN NO EVENT, SHALL DANIEL BE LIABLE FOR ANY INCIDENTAL, PUNITIVE, SPECIAL OR CONSEQUENTIAL DAMAGES INCLUDING, BUT NOT LIMITED TO, LOSS OF PRODUCTION, LOSS OF PROFITS, LOSS OF REVENUE OR USE AND COSTS INCURRED INCLUDING WITHOUT LIMITATION FOR CAPITAL, FUEL AND POWER, AND CLAIMS OF THIRD PARTIES.

PRODUCT NAMES USED HEREIN ARE FOR MANUFACTURER OR SUPPLIER IDENTIFICATION ONLY AND MAY BE TRADEMARKS/REGISTERED TRADEMARKS OF THESE COMPANIES.

Page 8 of the Emerson 3812 user manual — Warranty and Limitations
Page 8 · click to enlarge
Page 8

Warranty and Limitations

1. LIMITED WARRANTY: Subject to the limitations contained in Section 2 herein, Daniel Measurement & Control, Inc. ("Daniel") warrants that the licensed firmware embodied in the Goods will execute the programming instructions provided by Daniel, and that the Goods manufactured by Daniel will be free from defects in materials or workmanship under normal use and care and Services will be performed by trained personnel using proper equipment and instrumentation for the particular Service provided. The foregoing warranties will apply until the expiration of the applicable warranty period. Goods are warranted for twelve (12) months from the date of initial installation or eighteen (18) months from the date of shipment by Daniel, whichever period expires first. Consumables and Services are warranted for a period of 90 days from the date of shipment or completion of the Services. Products purchased by Daniel from a third party for resale to Buyer ("Resale Products") shall carry only the warranty extended by the original manufacturer. Buyer agrees that Daniel has no liability for Resale Products beyond making a reasonable commercial effort to arrange for procurement and shipping of the Resale Products. If Buyer discovers any warranty defects and notifies Daniel thereof in writing during the applicable warranty period, Daniel shall, at its option, correct any errors that are found by Daniel in the firmware or Services or repair or replace F.O.B. point of manufacture that portion of the Goods or firmware found by Daniel to be defective, or refund the purchase price of the defective portion of the Goods/Services. All replacements or repairs necessitated by inadequate maintenance, normal wear and usage, unsuitable power sources or environmental conditions, accident, misuse, improper installation, modification, repair, use of unauthorized replacement parts, storage or handling, or any other cause not the fault of Daniel are not covered by this limited warranty, and shall be at Buyer's expense. Daniel shall not be obligated to pay any costs or charges incurred by Buyer or any other party except as may be agreed upon in writing in advance by Daniel. All costs of dismantling, reinstallation and freight and the time and expenses of Daniel's personnel and representatives for site travel and diagnosis under this warranty clause shall be borne by Buyer unless accepted in writing by Daniel. Goods repaired and parts replaced by Daniel during the warranty period shall be in warranty for the remainder of the original warranty period or ninety (90) days, whichever is longer. This limited warranty is the only warranty made by Daniel and can be amended only in a writing signed by Daniel. THE WARRANTIES AND REMEDIES SET FORTH ABOVE ARE EXCLUSIVE. THERE ARE NO REPRESENTATIONS OR WARRANTIES OF ANY KIND, EXPRESS OR IMPLIED, AS TO MERCHANTABILITY, FITNESS FOR PARTICULAR PURPOSE OR ANY OTHER MATTER WITH RESPECT TO ANY OF THE GOODS OR SERVICES. Buyer acknowledges and agrees that corrosion or erosion of materials is not covered by this warranty.

2. LIMITATION OF REMEDY AND LIABILITY: DANIEL SHALL NOT BE LIABLE FOR DAMAGES CAUSED BY DELAY IN PERFORMANCE. THE REMEDIES OF BUYER SET FORTH IN THIS AGREEMENT ARE EXCLUSIVE. IN NO EVENT, REGARDLESS OF THE FORM OF THE CLAIM OR CAUSE OF ACTION (WHETHER BASED IN CONTRACT, INFRINGEMENT, NEGLIGENCE, STRICT LIABILITY, OTHER TORT OR OTHERWISE), SHALL DANIEL'S LIABILITY TO BUYER AND/OR ITS CUSTOMERS EXCEED THE PRICE TO BUYER OF THE SPECIFIC GOODS MANUFACTURED OR SERVICES PROVIDED BY DANIEL GIVING RISE TO THE CLAIM OR CAUSE OF ACTION. BUYER AGREES THAT IN NO EVENT SHALL DANIEL'S LIABILITY TO BUYER AND/OR ITS CUSTOMERS EXTEND TO INCLUDE INCIDENTAL, CONSEQUENTIAL OR PUNITIVE DAMAGES. THE TERM "CONSEQUENTIAL DAMAGES" SHALL INCLUDE, BUT NOT BE LIMITED TO, LOSS OF ANTICIPATED PROFITS, REVENUE OR USE AND COSTS INCURRED INCLUDING WITHOUT LIMITATION FOR CAPITAL, FUEL AND POWER, AND CLAIMS OF BUYER'S CUSTOMERS.

Page 9 of the Emerson 3812 user manual — Table of Contents
Page 9 · Contents
Contents

Contents

DanielTM 3812 Liquid Ultrasonic Flow Meter. Installation Manualp.1Contentsp.9Prefacep.3Daniel customer servicep.3Signal words and symbolsp.4Important safety instructionsp.5Noticep.7Warranty and Limitationsp.8List of Tablesp.11List of Figuresp.13Section 1: Introductionp.151.1 Typical Applicationsp.151.2 Features and benefitsp.161.3 Acronyms, abbreviations and definitionsp.171.4 Daniel MeterLink softwarep.191.5 Daniel 3812 Liquid Ultrasonic Flow meter designp.201.6 Meter specificationsp.271.7 Pre-installation considerationsp.311.8 Safetyp.311.9 Daniel 3812 Certifications and Approvalsp.331.10 FCC compliancep.33Section 2: Mechanical installationp.352.1 Meter piping, lifting and mountingp.352.2 Piping recommendationsp.422.3 Meter safety for hoist rings and lifting slingsp.462.3.1 Use of appropriate safety engineered swivel hoist rings in meter end flangesp.472.3.2 Appropriately rated lifting slingsp.542.4 Mounting requirements in heated or cooled pipelinesp.57Section 3: Electrical installationp.593.1 Cable length TTL modep.593.2 Cable length Open Collector modep.593.3 Grounding meter electronics housingp.603.4 Conduit sealsp.613.4.1 Startup for systems using explosion-proof conduitp.623.4.2 Startup for systems that use flame-proof cablep.633.5 Wiring and I/Op.643.5.1 CPU Module labeling and LED indicatorsp.653.6 I/O connectionsp.693.6.1 Frequency/Digital outputsp.703.6.2 Analog input settingsp.743.6.3 Analog output settingsp.743.6.4 Digital Inputp.743.6.5 DHCP server switch settingsp.743.6.6 Configuration protect switch settingsp.753.6.7 External power source connection and fusep.753.7 Security seal installationp.763.7.1 Direct or remote mount transmitter electronics enclosure sealp.763.7.2 Base Enclosure Security Sealsp.783.7.3 Seal the Junction Box remote mount electronics optionp.793.7.4 Bolted band shroud security sealsp.803.7.5 Clamped band shroud security sealsp.813.7.6 Split shroud security sealsp.823.7.7 Latched band shroud security sealsp.833.7.8 Seal conduit portsp.85Section 4: Configurationp.874.1 Daniel MeterLink setupp.874.2 Field Setup Wizardp.884.3 Using AMS Device Manager to configure the meterp.924.4 Using a Field Communicator to configure the meterp.1074.5 Security seals for the meterp.110Appendix A: Engineering drawingsp.111A.1 Daniel 3812 Liquid Ultrasonic Flow Meter drawingsp.111Appendix B: Open source licensesp.117B.1 GNU General Public Licensep.118B.2 GNU Lesser General Public Licensep.129B.3 BSD Open Source Licensep.133B.4 M.I.T Licensep.134Appendix C Indexp.135C.1 Manual Indexp.135
Page 11 of the Emerson 3812 user manual — List of Tables
Page 11 · click to enlarge
Page 11

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual List of Tables 3-9000-765 Rev H June 2015

List of Tables

Table 1-1 Acronyms, abbreviations and definitions 3

Table 1-2 Meter specifications 13

Table 2-1 3812 Ultrasonic Meter shrouds options per ANSI pressure rating 27

Table 2-2 Piping recommendation for uni-directional or bi-directional flow 31

Table 2-3 Hoist ring part number lookup table 39

Table 2-4 Hoist ring lookup table for Daniel Model 3812 Liquid Ultrasonic Flow Meters 39

Table 3-1 Configurations for open collector frequency outputs 45

Table 3-2 CPU Module labeling and LED functions 52

Table 3-3 Ethernet cable to PC communication 53

Table 3-4 Serial Port A parameters 54

Table 3-5 Frequency/Digital Outputs possible configurations 58

Table 3-6 DHCP server switch settings 60

Table 3-7 Configuration protect switch settings 61

Table 4-1 Local display labels, descriptions and valid units 75

Table B-1 Open source licences 99

List of Tables i

Page 12 of the Emerson 3812 user manual — List of Tables
Page 12 · click to enlarge
Page 12

List of Tables Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual June 2015 3-9000-765 Rev H

ii List of Tables

Page 13 of the Emerson 3812 user manual — List of Figures
Page 13 · click to enlarge
Page 13

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual List of Figures 3-9000-765 Rev H June 2015

List of Figures

Figure 1-1 Daniel MeterLink download and registration 5

Figure 1-2 Direct mount electronics assembly with split shroud 6

Figure 1-3 Direct mount electronics with latched single band shrouds and remote display 7

Figure 1-4 Direct mount electronics with bolted single band shrouds and local display 8

Figure 1-5 Direct mount electronics assembly with clamped band shrouds 9

Figure 1-6 Remote mount electronics assembly with clamped band shrouds 10

Figure 1-7 Remote mount electronics assembly with split shrouds 11

Figure 1-8 Optional local display and glass endcap 12

Figure 2-1 Direct mount meter electronics assembly with split shroud 24

Figure 2-2 Remote mount meter electronics assembly with split shroud 25

Figure 2-3 Direct mount meter electronics with bolted band shrouds 26

Figure 2-4 Direct mount meter electronics assembly with clamped band shrouds 27

Figure 2-5 Piping recommendations unidirectional flow 29

Figure 2-6 Piping recommendations bidirectional flow 30

Figure 2-7 Meter end flange with tapped flat-counterbore hole for hoist ring 33

Figure 2-8 Safety approved hoist ring and non-compliant eye bolt 34

Figure 2-9 90 Degree angle between slings 35

Figure 2-10 Sling contacting electronics enclosure 36

Figure 2-11 Correct sling attachment 41

Figure 2-12 Incorrect sling attachment 42

Figure 3-1 Transmitter Electronics Enclosure internal chassis ground 46

Figure 3-2 External ground lug 47

Figure 3-3 CPU Module labeling and LED indicators 51

Figure 3-4 PC to meter serial connection wiring 55

Figure 3-5 CPU Module I/O connections 55

Figure 3-6 CPU Module - Frequency/Digital outputs common ground 59

Figure 3-7 CPU Module power source connections 61

Figure 3-8 Transmitter electronics enclosure security latch 62

Figure 3-9 Direct or remote mount Transmitter Electronics Enclosure security seals 63

Figure 3-10 Base Enclosure security seals 64

Figure 3-11 3812 Remote mount transmitter electronics option 65

Figure 3-12 Bolted band shroud security seals - bottom view 66

Figure 3-13 Clamped shroud security seals 67

List of Figures iii

Page 14 of the Emerson 3812 user manual — List of Figures
Page 14 · click to enlarge
Page 14

List of Figures Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Installation Manual June 2015 3-9000-765 Rev H

Figure 3-14 Split shroud security seals 68

Figure 3-15 Latched band shroud assembly 69

Figure 3-16 Shroud latch holes for security wire seals 70

Figure 4-1 AMS Device Description search 78

Figure 4-2 AMS file download complete 79

Figure 4-3 AMS Device Manager 80

Figure 4-4 AMS Device Manager - Overview 80

Figure 4-5 AMS Device Manager - Guided Setup 81

Figure 4-6 AMS Device Manager - Service Tools All Variables status indicators 83

Figure 4-7 Display Meter K-Factors 83

Figure 4-8 AMS Device Manager - Configure Manual Setup 84

Figure 4-9 Gating configuration parameter Edge gated, active high 86

Figure 4-10 Gating configuration parameter Edge gated, active low 86

Figure 4-11 Gating configuration parameter State gated, active high 86

Figure 4-12 Gating configuration parameter State gated, active low 87

Figure 4-13 Configure Flow Analysis Alert 87

Figure 4-14 AMS Device Manager - Service Tools Alerts 89

Figure 4-15 Configuration changes dialog 89

Figure 4-16 AMS Device Manager - Service Tools 90

Figure 4-17 AMS Device Manager - Service Tools All Variables 91

Figure 4-18 AMS Device Manager - Service Tools Trends 92

Figure 4-19 3812 transmitter field wiring conduit entries 94

Figure 4-20 Field Communicator wiring diagram for the 3810 Series electronics 95

iv List of Figures

Page 15 of the Emerson 3812 user manual — 1.1 Typical Applications
Page 15 · click to enlarge
Page 15

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 1: Introduction 3-9000-765 Rev H May 2015

Section 1: Introduction

Daniel 3812 Liquid Ultrasonic Flow Meters have direct mount or remote mount electronic options and various configurations that meet a broad range of customer requirements. Each meter comes fully assembled from Daniel Measurement and Control, Inc. and all parts and assemblies are tested prior to shipment. Refer to the following documents for additional details:

  • P/N 3-9000-761 HART® Field Device Specification for Liquid Ultrasonic Flow Meters
  • P/N 3-9000-763 Daniel MeterLink Software for Daniel Gas and Liquid Ultrasonic Flow Meters
  • P/N 3-9000-767 Daniel Model 3812 Liquid Ultrasonic Flow Meter Maintenance and Troubleshooting Manual

The Daniel 3812 Liquid Ultrasonic Flow Meter technology can be applied to allocation measurement and check metering applications as shown below.

1.1 Typical Applications

  • Allocation measurement
  • Check metering
  • Leak detection
  • Line balancing
  • Batch control
  • Loading and off loading
  • Offshore

- FPSO (Floating Production, Storage and Offshore Loading) - Offshore Platforms - Barges

  • Pipelines

- Crude Oil pipelines - Refined product pipelines

  • Terminals

- Loading and off-loading (Ship, barge, truck, railcar, etc…) - Tank Farms

  • Cavern Storage

Typical Applications 1

Page 16 of the Emerson 3812 user manual — 1.2 Features and benefits
Page 16 · click to enlarge
Page 16
Section 1Introduction Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

1.2 Features and benefits

  • Explosion-proof transmitter electronics enclosure with CPU Module, Power Supply, Intrinsic Safety Barrier Module
  • Intrinsically safe transducer electronics enclosure with the Acquisition Module
  • Daniel MeterLink (software for Daniel Ultrasonic Flow Meters)
  • HART ® and AMS Suite: Intelligent Device Manager communications for PlantWeb TM architecture
  • Direct mount or remote mount options electronics
  • Reduce unaccounted measurement
  • Increase energy savings
  • Replaceable transducers while under pressure
  • Extensive self diagnostics
  • Immediate alarm reporting
  • Auto-detected ASCII/RTU Modbus communications protocol
  • Interchangeable electronics modules
  • Internet-ready communications
  • Ethernet access
  • Modbus TCP/IP
  • On-board LED status indicators
  • Analog pressure and temperature inputs
  • Local display and glass endcap (optional)

For other features and benefits refer to the product datasheet at

http://www2.emersonprocess.com/EN-US/BRANDS/DANIEL/FLOW/Pages/Flow.aspx

2 Features and benefits

Page 17 of the Emerson 3812 user manual — 1.3 Acronyms, abbreviations and definitions
Page 17 · click to enlarge
Page 17

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 1: Introduction 3-9000-765 Rev H May 2015

1.3 Acronyms, abbreviations and definitions

Table 1-1 Acronyms, abbreviations and definitions

Acronym or abbreviation Definition

oC degrees celsius (temperature unit)

oF degrees fahrenheit (temperature unit)

ADC analog-to-digital converter AI analog input AMS® Device Manager Asset Management Software - Device Manager AO analog output ASCII MODBUS A Modbus protocol message framing format in which ASCII characters are used to delineate the beginning and end of the frame. ASCII stands for American Standard Code for Information Interchange. boolean a type of data point that can only take on values of TRUE or FALSE (generally TRUE is represented by a value of 1, FALSE is represented by a value of 0) bps bits per second (baud rate) cPoise centipoise (viscosity unit) CPU central processing unit CTS Clear-to-Send; the RS-232C handshaking signal input to a transmitter indicating that it is okay to transmit data – i.e., the corresponding receiver is ready to receive data. Generally, the Request-to-Send (RTS) output from a receiver is input to the Clear-to- Send (CTS) input of a transmitter. DAC Digital-to-Analog Converter

Daniel MeterLinkTM Daniel ultrasonic meter interface software

DI digital input Direct Mount Transmitter electronics enclosure and base electronics enclosure is directly mounted to meter body DO digital output DHCP Dynamic Host Configuration Protocol dm decimeter (10-1 meters, length unit) ECC Error Correction Code EEPROM Electrically-Erasable, Programmable Read-Only Memory Flash non-volatile, programmable read-only memory FODO output that is user configurable as either a frequency or digital output HART® Communication Protocol Highway Addressable Remote Transducer communications protocol hr hour (time unit) Hz Hertz (cycles per second, frequency unit) I/O Input/Output IS Intrinsically Safe K Kelvin (temperature unit) kHz kilohertz (103 cycles per second, frequency unit) LAN Local Area Network LED light-emitting diode m meter (length unit)

m3/d cubic meters per day (volumetric flow rate)

Acronyms, abbreviations and definitions 3

° degree (angle)

Page 18 of the Emerson 3812 user manual — 1.3 Acronyms, abbreviations and definitions
Page 18 · click to enlarge
Page 18
Section 1Introduction Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

Table 1-1 Acronyms, abbreviations and definitions

Acronym or abbreviation Definition

m3/h cubic meters per hour (volumetric flow rate)

m3/s cubic meters per second (volumetric flow rate)

mA milliamp (current unit) MAC Address Media Access Control (Ethernet Hardware Address -EHA)

microinch (μinch) microinch (10-6 in)

micron micrometer (10-6 m) MMU Memory Management Unit MPa Megapascal (equivalent to 106 Pascal) (pressure unit) N/A not applicable

Nm3/h normal cubic meters per hour

NOVRAM non-volatile random access memory Pa Pascal, equivalent to 1 newton per square meter (pressure unit) Pa⋅s Pascal Second (viscosity unit) PC Personal Computer PFC peripheral field connection (board) P/N part number PS power supply (board) psi pounds per square inch (pressure unit) psia pounds per square inch absolute (pressure unit) psig pounds per square inch gage (pressure unit) R Radius rad radian (angle) RAM Random Access Memory Remote Mount Option Detaching the transmitter electronics enclosure and base electronics enclosure from meter body and mounting and affixing them with the mounting bracket to a pole or other structure. RTS Request-to-Send; the RS-232C handshaking signal output by a receiver when it is ready to receive data RTU MODBUS A Modbus protocol framing format in which elapsed time between received charac ters is used to separate messages. RTU stands for Remote Terminal Unit. s second (time unit, metric) SDRAM Synchronous Dynamic Random Access Memory sec second (time unit, U.S. Customary) TCP/IP Transmission Control Protocol/Internet Protocol time_t seconds since Epoch (00:00:00 UTC Jan. 1, 1970) (time unit) UDP User Datagram Protocol U.L. Underwriters Laboratories, Inc. - product safety testing and certification organization V volts (electric potential unit) W watts (power unit)

4 Acronyms, abbreviations and definitions

Page 19 of the Emerson 3812 user manual — 1.4 Daniel MeterLink software
Page 19 · click to enlarge
Page 19

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 1: Introduction 3-9000-765 Rev H May 2015

1.4 Daniel MeterLink software

Daniel MeterLink software has robust features for setting communications parameters, calibrating your meter, collecting logs and reports and monitoring the meter health and alarm statuses. Daniel MeterLink may be downloaded at no charge from:

http://www2.emersonprocess.com/en-US/brands/daniel/Flow/ultrasonics/Pages/MeterLink.aspx

Figure 1-1 Daniel MeterLink download and registration

Emerson 3812 — 1.4 Daniel MeterLink software diagram (page 19)
Fig. 3 · 1.4 Daniel MeterLink software
1. From the right panel under Quick Links, click the MeterLink Registration and Download link.
2. Click the Order Now button to complete the Online registration form.
3. Click Next to go to the order confirmation page.
4. Click Complete Order. You will receive a conformation email with a hyperlink directing you to the download site. Click the link provided.
5. Click Save.

Refer to the Daniel MeterLink Software for Gas and Liquid Ultrasonic Meters Quick Start Manual (P/N 3-9000-763) for installation instructions and setup for initial communications. You may download the manual from the Daniel MeterLink web page:

http://www2.emersonprocess.com/en-US/brands/daniel/Flow/ultrasonics/Pages/MeterLink.aspx

Daniel MeterLink software 5

Page 20 of the Emerson 3812 user manual — 1.5 Daniel 3812 Liquid Ultrasonic Flow meter design
Page 20 · click to enlarge
Page 20
Section 1Introduction Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

1.5 Daniel 3812 Liquid Ultrasonic Flow meter design

TheDaniel 3812 Liquid Ultrasonic Flow Meter designs include a direct or remote mount electronics option and depending on the meter’s outside diameter, have a shroud cover protecting the transducers and cable assemblies. See Table 2-1 and the list below for meter body shroud types.

Shroud options are:

  • split shroud
  • bolted band shroud
  • latched single band shroud
  • clamped band shroud

Figure 1-2 Direct mount electronics assembly with split shroud

Emerson 3812 — 1.5 Daniel 3812 Liquid Ultrasonic Flow meter design diagram (page 20)
Fig. 4 · 1.5 Daniel 3812 Liquid Ultrasonic Flow meter design

A.

B.

A. Explosion-proof transmitter enclosure (CPU Module, Power Supply, I.S. Barrier Board,

Backplane board, and optional LCD Board with glass endcap)

B. Intrinsically-safe base enclosure includes Acquisition Module C. Meter - body and split shroud for transducers and cables assemblies

6 Daniel 3812 Liquid Ultrasonic Flow meter design

Page 21 of the Emerson 3812 user manual — 1.5 Daniel 3812 Liquid Ultrasonic Flow meter design
Page 21 · click to enlarge
Page 21

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 1: Introduction 3-9000-765 Rev H May 2015

Figure 1-3 Direct mount electronics with latched single band shrouds and remote display

Emerson 3812 — 1.5 Daniel 3812 Liquid Ultrasonic Flow meter design diagram (page 21)
Fig. 5 · 1.5 Daniel 3812 Liquid Ultrasonic Flow meter design

A.

B.

A. Explosion-proof transmitter enclosure (CPU Module, Power Supply, I.S. Barrier Board,

Backplane board, and optional LCD Board with glass endcap)

B. Intrinsically-safe base enclosure includes Acquisition Module C. Meter - body and latched band shrouds for transducers and cable assemblies

Daniel 3812 Liquid Ultrasonic Flow meter design 7

Page 22 of the Emerson 3812 user manual — 1.5 Daniel 3812 Liquid Ultrasonic Flow meter design
Page 22 · click to enlarge
Page 22
Section 1Introduction Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

Figure 1-4 Direct mount electronics with bolted single band shrouds and local display

Emerson 3812 — 1.5 Daniel 3812 Liquid Ultrasonic Flow meter design diagram (page 22)
Fig. 6 · 1.5 Daniel 3812 Liquid Ultrasonic Flow meter design

A.

B.

A. Explosion-proof transmitter enclosure (CPU Module, Power Supply, I.S. Barrier Board,

Backplane board, and optional LCD Board with glass endcap)

B. Intrinsically-safe base enclosure includes Acquisition Module C. Meter - body and latched band shrouds for transducers and cable assemblies

8 Daniel 3812 Liquid Ultrasonic Flow meter design

Page 23 of the Emerson 3812 user manual — 1.5 Daniel 3812 Liquid Ultrasonic Flow meter design
Page 23 · click to enlarge
Page 23

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 1: Introduction 3-9000-765 Rev H May 2015

Figure 1-5 Direct mount electronics assembly with clamped band shrouds

Emerson 3812 — 1.5 Daniel 3812 Liquid Ultrasonic Flow meter design diagram (page 23)
Fig. 7 · 1.5 Daniel 3812 Liquid Ultrasonic Flow meter design

A.

B.

A. Explosion-proof transmitter enclosure (CPU Module, Power Supply, I.S. Barrier Board,

Backplane board, optional LCD Board with glass endcap)

B. Intrinsically-safe base enclosure includes Acquisition Module C. Meter - body and band shrouds for transducers and cable assemblies

Daniel 3812 Liquid Ultrasonic Flow meter design 9

Page 24 of the Emerson 3812 user manual — 1.5 Daniel 3812 Liquid Ultrasonic Flow meter design
Page 24 · click to enlarge
Page 24
Section 1Introduction Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

Figure 1-6 Remote mount electronics assembly with clamped band shrouds

Emerson 3812 — 1.5 Daniel 3812 Liquid Ultrasonic Flow meter design diagram (page 24)
Fig. 8 · 1.5 Daniel 3812 Liquid Ultrasonic Flow meter design

A.

B.

A. Explosion-proof transmitter enclosure (CPU Module, Power Supply, I.S. Barrier Board,

Backplane board, and optional LCD Board with glass endcap)

B. Intrinsically-safe base enclosure includes Acquisition Module C. Meter - body and band shrouds for transducers and cable assemblies

10 Daniel 3812 Liquid Ultrasonic Flow meter design

Page 25 of the Emerson 3812 user manual — 1.5 Daniel 3812 Liquid Ultrasonic Flow meter design
Page 25 · click to enlarge
Page 25

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 1: Introduction 3-9000-765 Rev H May 2015

Figure 1-7 Remote mount electronics assembly with split shrouds

Emerson 3812 — 1.5 Daniel 3812 Liquid Ultrasonic Flow meter design diagram (page 25)
Fig. 9 · 1.5 Daniel 3812 Liquid Ultrasonic Flow meter design

A.

B.

D.

E.

A. Explosion-proof transmitter enclosure (CPU Module, Power Supply, I.S. Barrier Board

Backplane board, optional LCD Board with glass endcap)

B. Intrinsically-safe base enclosure includes Acquisition Module C. Acquisition cable conduit

D. Junction box E. Meter - body and split shrouds for transducer and cable assemblies

Daniel 3812 Liquid Ultrasonic Flow meter design 11

Page 26 of the Emerson 3812 user manual — 1.5 Daniel 3812 Liquid Ultrasonic Flow meter design
Page 26 · click to enlarge
Page 26
Section 1Introduction Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

Figure 1-8 Optional local display and glass endcap

Emerson 3812 — 1.5 Daniel 3812 Liquid Ultrasonic Flow meter design diagram (page 26)
Fig. 10 · 1.5 Daniel 3812 Liquid Ultrasonic Flow meter design

The Daniel 3812 Liquid Ultrasonic Flow Meter is a two-path (four transducers) in-line meter designed to measure the difference in signal transit time with and against the flow across one or more measurement path(s). A signal transmitted in the flow direction travels faster than one transmitted against the flow direction. Each measurement path is defined by a transducer pair in which each transducer alternately acts as transmitter and receiver. The meter uses transit time measurements and transducer location information to calculate the mean velocity.

Computer simulations of various velocity profiles demonstrate that multiple measurement paths provide an optimum solution for measuring asymmetric flow. The Daniel 3812 Liquid Ultrasonic Flow Meter utilizes two cross-bore, parallel-plane measurement paths, offers a high degree of repeatability, bi-directional measurement and superior low-flow capabilities without the compromises associated with conventional technologies.

The Daniel 3812 Liquid Ultrasonic Flow Meter’s U.L. safety listing is accomplished through the combination of an explosion-proof Transmitter Electronics Enclosure that houses the CPU Module, Power Supply board, I.S. Barrier board, Backplane board and optional LCD Display board. NOTE: The optional LCD Display requires firmware v1.04 and Uboot version, January 31, 2013.

The Base Electronics Enclosure is intrinsically safe and houses the Acquisition Module, the acquisition cable and wiring. The Intrinsically safe transducers and cable assemblies are designed for Class 1, Division 1, Groups C and D areas without need of further protection when installed in accordance with the field wiring diagram (refer to Daniel drawing DMC - 004936, see Appendix A and Section 1.8).

12 Daniel 3812 Liquid Ultrasonic Flow meter design

Page 27 of the Emerson 3812 user manual — ESCAPING FLUIDS HAZARD
Page 27 · click to enlarge
Page 27

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 1: Introduction 3-9000-765 Rev H May 2015

1.6 Meter specifications

ESCAPING FLUIDS HAZARD

The purchaser of the meter is responsible for the selection of Daniel components/seals and materials compatible with the chemical properties of the measurement fluid.

Failure to select suitable meter components/seals may cause escaping fluids, resulting in injury or equipment damage.

Consult your Daniel Sales and Service representative to ensure you purchase the correct components and seals for your application. Table 1-2 Meter specifications

Liquid Ultrasonic Flow Meter specifications

  • Meter type Number of paths
  • Two path (four transducer) chordal design Enclosure materials
  • ASTM B26 Gr A356.0 T6 Aluminum

— Chromate conversion coated with a polyurethane enamel • ASTM A351 Gr CF8M Stainless Steel

  • — Passivated Ultrasonic type
  • Transit-time based measurement
  • Spool piece with integral mount transducers Enclosure materials
  • ASTM B26 Gr A356.0 T6 Aluminum

— Chromate conversion coated with a polyurethane enamel • ASTM A351 Gr CF8M Stainless Steel

  • Linearity
  • ± 0.30% of measured value over a 10:1 turndown (40 to 4 ft/s; 12.2 to 1.2 m/s) Repeatability
  • ±0.10% of reading in the specified velocity range Velocity range
  • 40.0 ft/s (12.2 m/s) (nominal) to 2.0 ft/s (0.6 m/s)
  • 48 fps (14.3 m/s) (over-range) Body and Flange Pressure rating range
  • U.S. Customary Units sizes - 2, 3, 4, 6, 8, 10, 12, 14, 16, 18, 20, 24, 28, 30, and 36 (in)
  • ANSI pressure classes 150 and 300, 600, 900 (per ANSI B16.5)
  • Carbon Steel
  • 316 Stainless Steel Metric Units sizes
  • DN 50, 75, 100, 150, 200, 250, 300, 350, 400, 450, 500, 600, 700, 750, and 900
  • PN 20, 50, 100, 150
  • Carbon Steel
  • 316 Stainless Steel Meter bore
  • Schedule 40 and Schedule 80 Maximum Pressures
  • Dependent on operating temperature

— Passivated · Meter Performance · Meter specifications 13

Page 28 of the Emerson 3812 user manual — 1.6 Meter specifications
Page 28 · click to enlarge
Page 28
Section 1Introduction Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

Table 1-2 Meter specifications

  • Flange types ANSI classes - 150, 300, 600 and 900
  • Raised face or RTJ Specific Gravity 0.35 to 1.50 Accuracy Limits Accuracy limits typically are:
  • ± 2% without a flow calibration, ± 0.3% with flow calibration Minimum operating pressure
  • 0 psig
  • 0 barg

Power Meter • 10.4 VDC to 36 VDC measured at the meter terminals1

  • 11 W maximum power consumption Serial cable
  • Belden #9940 or equivalent (22 gauge)
  • — Capacitance (pF/m) 121.397 (conductor to conductor) — Capacitance (pF/m) 219.827 (conductor to other conductor and shield) — Resistance (DC) DCR @ 20 °C (Ohm/km) 48.2307 Nominal Outer shield resistance - DCR @ 20°C (Ohm/km) 16.405 — Operating voltage - 300 V RMS (UL AWM Style 2464) — Current 2.4 Amps per conductor @ 25°C (recommended) Ethernet cable
  • Cat-5 Standard 100Mbps Frequency ( see Table 3-1 )
  • 22 AWG wire characteristics are as follows: — Capacitance = 20 pF/ft or 20 nF/1000 ft (between two wires) — Resistance = 0.0168 Ohms/ft or 16.8 Ohms/1000 ft — Pull-up voltage is 24 VDC Temperature flameproof transmitter electronic enclosure and base electronic enclosure
  • Ambient: -40 o F to 140 o F (-40 o C to 60 o C)
  • Storage: -58 o F to 185 o F (-50 o C to 85 o C)

operating temperature exceeds 140 oF (60 oC). Transducers • LT-10 and LT-11 Operating temperature range with NBR O-rings

-58 oF to +275 oF (-50 oC to 135 oC) • LT-10 and LT-11 Operating temperature range with FKM O-rings

-40 oF to +302 oF (-40 oC to +150 oC)

the transducers. Note: LT-10 transducers are designed for 4 inch to 10 inch meters. LT-11 transducers are designed

for 12 inch and larger meters. Note: The ultrasonic transducers are not intended for use across boundary walls of different

hazardous area classifications. The transmitter electronics cannot be remote mounted from a Division 1 classification to a Division 2 area to meet an area classification. Acquisition Cable • Total cable length between the Acquisition Module and ultrasonic transducers must not exceed 15 feet (4.7 meters) when using the remote mount option (see Figure 1-7)

Meter Performance (continued) · Electronic specifications · 14 Meter specifications

Page 29 of the Emerson 3812 user manual — 1.6 Meter specifications
Page 29 · click to enlarge
Page 29

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 1: Introduction 3-9000-765 Rev H May 2015

Table 1-2 Meter specifications

  • Connectivity protocols One serial RS-232/RS-485 port (115 kbps baud rate) (Modbus RTU/ASCII)
  • (1) Serial Port A (RS-232/RS-485 Full Duplex/RS-485 Half Duplex) One Ethernet Port (TCP/IP) 100 BaseT
  • Modbus TCP Device compatibility FloBoss 103, FloBoss S600 flow computer, ROC 107

Digital, analog, and frequency inputs

  • (1) Single polarity (for flow calibration gating - contact closure)
  • Single input for starting and stopping
  • Four pulse configurations available Analog Input(s) (2) 4-20 mA
  • AI-1 Temperature
  • AI-2 Pressure Note: The analog-to-digital conversion accuracy is within ±0.05% of full scale over the operating

temperature range.

series resistance so HART® Communicators can be connected to configure sensors. A 24 Volt DC power supply is available to provide power to the sensors.

Communications specifications · Digital Input(s) (Selectable) · Meter specifications 15

Page 30 of the Emerson 3812 user manual — 1.6 Meter specifications
Page 30 · click to enlarge
Page 30
Section 1Introduction Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

Table 1-2 Meter specifications

Digital, analog, and frequency outputs

  • The meter has user-configurable selections for either a frequency output or Digital status (FODO) (Also see Section 3.6.1 ) (3) Frequency/Digital Outputs
  • FODO1 (four possible output configurations)
  • FODO2(eight possible output configurations)
  • FODO3(eight possible output configurations)
  • Frequency or Digital Output parameter pairs ( see Section 3.6.1 ) Frequency or Digital Outputs (FODO 1) source selections:
  • (FO1A, DO1A, FO1B, DO1B) Frequency or Digital Outputs (FODO 2) source selections
  • (FO1A, DO1A, FO1B, DO1B, FO2A, DO2A, FO2B, DO2B) Frequency or Digital Outputs (FODO 3) source selections
  • (FO1A, DO1A, FO1B, DO1B, FO2A, DO2A, FO2B, DO2B)
  • Mode options:
  • Open Collector (requires external excitation supply voltage and pull-up resistor)
  • TTL (internally powered by the meter 0-5 VDC signal)
  • Channel B Phase options:
  • Lag forward, Lead reverse (Phase B lags Phase A while reporting forward flow, leads Phase A while reporting reverse flow)
  • Lead forward, Lag reverse (Phase B leads Phase A while reporting forward flow, lags Phase A while reporting reverse flow)
  • Phase A and Phase B output (based on flow direction)
  • Reverse flow - output only reports flow in the reverse direction. For frequency outputs, Phase B of the output is 90 degrees out of phase with Phase A.
  • Forward flow - output only reports flow in the forward direction. For frequency outputs, Phase B of the output is 90 degrees out of phase with Phase A.
  • Absolute - output reports flow in both directions. For frequency outputs, Phase B of the output is 90 degrees out of phase with Phase A.
  • Bidirectional - output reports flow on Phase A only in the forward direction and on Phase B only in the reverse direction.
  • Maximum frequency for the frequency outputs
  • 1000Hz
  • 5000Hz Analog Output(s)
  • (1) 4-20 mA independently configurable analog output (HART) (1) 4-20 mA independently configurable analog output (conventional) The analog output zero scale offset error is within ±0.1% of full scale and gain error is within ±0.2% of full scale. The total output drift is within ±50 ppm of full scale per °C.

Frequency/Digital Output(s) · 16 Meter specifications

Page 31 of the Emerson 3812 user manual — DANGER TO PERSONNEL AND EQUIPMENT
Page 31 · click to enlarge
Page 31

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 1: Introduction 3-9000-765 Rev H May 2015

1.7 Pre-installation considerations

  • Pipeline equipment code compliance, ANSI, ASME, etc.
  • Proper Inlet/outlet meter tube piping for reasonable stable flow to the settling chamber (first meter tube spool upstream of the meter).
  • Electrical safety compliance; UL, CSA, ATEX, IECEx etc.
  • Civil and structural good practices compliance
  • Contractual agreements or governmental compliance (or both)
  • In-situ performance test procedures
  • Field tested advanced meter health and flow dynamics diagnostics
  • Data collection and retention procedures

1.8 Safety

The Daniel 3812 Liquid Ultrasonic Flow Meter is suitable for use in U.L. Class 1, Division 1, Group C and D hazardous locations.

DANGER TO PERSONNEL AND EQUIPMENT

Observe all precautionary labels posted on the equipment and safety messages throughout the meter documentation.

Failure to do so may result in injury to personnel or cause damage to the equipment.

Daniel 3810 Series Liquid Ultrasonic Meters are approved to the ATEX Directive 94/9/EC.

Refer to the 3810 Series Systems Wiring Diagram, Sheet 3 (P/N DMC -004936) for the certification tag (see Appendix A).

Pre-installation considerations 17

Page 32 of the Emerson 3812 user manual — 1.7 Pre-installation considerations
Page 32 · click to enlarge
Page 32
Section 1Introduction Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

Daniel 3810 Series Liquid Ultrasonic Meters are INMETRO certified. Refer to the 3810 Series Liquid Ultrasonic Flow Meter Tag, INMETRO Certification drawing DMC - 006173.

Certificate number: NCC 11.0163 X Marking: --Ex d ia IIB T4 Gb IP66 W Electrical parameters: Refer to Section 1.6, Table 1-2.

Special conditions for safe use

  • Explosion proof joint dimensions are compliant with the Brazilian Association of technical standard: ABNT NBR IEC 60079-1, Table 3.
  • The enclosure for the explosion proof transmitter and intrinsically safe barrier must be remote mounted (Refer to Section 1.6 , Table 1-2 ) if the operating temperature exceeds 1

40 oF (60 oC) (refer to Section 1.6, Table 1-2).

  • Cable length (refer to Section 1.6 , Table 1-2 ).

18 Safety

Page 33 of the Emerson 3812 user manual — 1.10 FCC compliance
Page 33 · click to enlarge
Page 33

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 1: Introduction 3-9000-765 Rev H May 2015

1.9 Daniel 3812 Certifications and Approvals

Daniel 3810 Series Liquid Ultrasonic Flow Meters have electrical, metrology, intrinsic safety and Pressure Equipment Directive certifications, approvals and lab testing and calibration certifications by the agencies listed below. Refer to the nameplate tag on the meter body, the wiring diagram (Drawing DMC - 0004936) in Appendix A and observe all safety precautions. Daniel 3810 Series Liquid Ultrasonic Flow Meters operate within the pressure and temperature range of the device (also see Section 1.6 for meter specifications).

The Daniel 3810 Series Liquid Ultrasonic Flow Meters certifications and approvals are available from Daniel Measurement and Control, Inc.

  • ATEX (British Approval Service for Electrical Equipment in Flammable Atmospheres)
  • IECEx (International Electrotechnical Commission for explosive atmospheres)
  • U.L. (Underwriter Laboratories)
  • U.L.C. (Underwriter Laboratories of Canada)
  • PED (BSI Group) British Standards Institution
  • INMETRO (National Institute of Metrology, Quality and Technology)
  • Demko (Dansk Elektrisk Materiel Kontrol - Danish Electrical Equipment Supervision)
  • GOST R (Government Standard Russia)
  • L.A.B. (Laboratory Accreditation Bureau) ISO/IEC 17025:2005

1.10 FCC compliance

This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment.

This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.

Changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment.

Daniel 3812 Certifications and Approvals 19

Page 34 of the Emerson 3812 user manual — 1.9 Daniel 3812 Certifications and Approvals
Page 34 · click to enlarge
Page 34
Section 1Introduction Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

20 FCC compliance

Page 35 of the Emerson 3812 user manual — TRIPPING HAZARD
Page 35 · click to enlarge
Page 35

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 2: Mechanical installation 3-9000-765 Rev H May 2015

Section 2: Mechanical installation

2.1 Meter piping, lifting and mounting

Refer to the following sections for piping recommendations, lifting with hoist rings and slings, mounting in heated or cooled pipelines and safety warnings and precautions.

SURFACE TEMPERATURE HAZARD

The meter body and piping may be extremely hot or cold.

Wear appropriate personal protective equipment when coming in contact with the meter. Failure to do so may result in injury.

CUTTING HAZARD

Sharp edges may be present on the meter.

Wear appropriate personal protective equipment when working on the meter. Failure to do so may cause serious injury

TRANSPORTATION HAZARD

When moving the meter, do not insert the forks of a forklift into the bore.

Inserting the forks may cause the meter to become unstable, resulting in injury or damage to the bore and sealing face.

TRIPPING HAZARD

Clear all obstacles or obstructions from the work area when transporting, installing or removing the meter.

Failure to clear the work area may cause injury to personnel.

Meter piping, lifting and mounting 21

Page 36 of the Emerson 3812 user manual — ESCAPING FLUIDS HAZARD
Page 36 · click to enlarge
Page 36

Section 2: Mechanical installation Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

CRUSHING HAZARD

Do not remove flange stabilizers.

Attempting to do so may allow the meter to roll, resulting in serious injury or equipment damage.

A.

A. Flange stabilizers

ESCAPING FLUIDS HAZARD

The purchaser of the meter is responsible for the selection of Daniel components/seals and materials compatible with the chemical properties of the measurement fluid.

Failure to select suitable meter components/seals may cause escaping fluids, resulting in injury or equipment damage.

Consult your Daniel Sales and Service representative to ensure you purchase the correct components and seals for your application.

22 Meter piping, lifting and mounting

Page 37 of the Emerson 3812 user manual — FLUID CONTENTS MAY BE HAZARDOUS
Page 37 · click to enlarge
Page 37

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 2: Mechanical installation 3-9000-765 Rev H May 2015

Meter components

FLUID CONTENTS MAY BE UNDER PRESSURE

When the meter is under pressure, DO NOT attempt to remove or adjust the transducer housing.

Attempting to do so may release pressurized fluid, resulting in serious injury or equipment damage.

FLUID CONTENTS MAY BE HAZARDOUS

The meter must be fully depressurized and drained before attempting to remove the transducer housing. If fluid begins to leak from the transducer housing, immediately reinstall it.

Failure to do so may cause serious injury or equipment damage.

Emerson 3812 — FLUID CONTENTS MAY BE HAZARDOUS diagram (page 37)
Fig. 11

A.

A. Transducer housing

Meter piping, lifting and mounting 23

Page 38 of the Emerson 3812 user manual — Section 2: Mechanical installation
Page 38 · click to enlarge
Page 38

Section 2: Mechanical installation Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

Daniel Liquid Ultrasonic Flow Meters are assembled, configured, and tested at the factory. The meter components include the transmitter electronics enclosure, the base electronics enclosure, the meter body with shroud covers for the transducers and cable assemblies and a direct or remote mount option.

Figure 2-1 Direct mount meter electronics assembly with split shroud

Emerson 3812 — Section 2: Mechanical installation diagram (page 38)
Fig. 12 · Section 2: Mechanical installation

A.

B.

A. Direct mount- explosion-proof transmitter enclosure (CPU Module, Power Supply,

I.S. Barrier Board, Backplane Board, and optional LCD Display Board with glass endcap)

B. Intrinsically-safe base enclosure includes Acquisition Module C. Meter - body and split shroud cover for transducers and cables assemblies

24 Meter piping, lifting and mounting

Page 39 of the Emerson 3812 user manual — Section 2: Mechanical installation
Page 39 · click to enlarge
Page 39

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 2: Mechanical installation 3-9000-765 Rev H May 2015

Figure 2-2 Remote mount meter electronics assembly with split shroud

A.

B.

D.

E.

A. Remote mount explosion-proof transmitter enclosure (CPU Module, Power Supply, I.S.

I.S. Barrier Board, Backplane Board and optional LCD Display Board)

B. Intrinsically-safe base enclosure includes Acquisition Module C. Acquisition cable conduit

D. Junction box with acquisition cable terminal blocks E. Meter - body and split shroud cover for transducer assemblies and cables

Meter piping, lifting and mounting 25

Page 40 of the Emerson 3812 user manual — Section 2: Mechanical installation
Page 40 · click to enlarge
Page 40

Section 2: Mechanical installation Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

Figure 2-3 Direct mount meter electronics with bolted band shrouds

A.

B.

A. Explosion-proof transmitter enclosure (CPU Module, Power Supply, I.S. Barrier Board,

Backplane Board and optional LCD Board with glass endcap)

B. Intrinsically-safe base enclosure includes Acquisition Module C. Meter - body and split shroud cover for transducers and cables assemblies

26 Meter piping, lifting and mounting

Page 41 of the Emerson 3812 user manual — Section 2: Mechanical installation
Page 41 · click to enlarge
Page 41

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 2: Mechanical installation 3-9000-765 Rev H May 2015

Figure 2-4 Direct mount meter electronics assembly with clamped band shrouds

A.

B.

A. Explosion-proof transmitter enclosure (CPU Module, Power Supply, I.S. Barrier Board

B. Intrinsically-safe base enclosure includes Acquisition Module C. Meter - body and split shroud cover for transducers and cables assemblies

Table 2-1 3812 Ultrasonic Meter shrouds options per ANSI pressure rating

Meter body size ANSI pressure rating Shroud type

2” - 3” 150 and 300 Bolted band shroud or latched band shroud

4” - 10” 150 and 300 Split shroud or latched band shroud

600 and 900 Clamped band shroud

12” and larger 150, 300, 600, 900 Clamped band shroud

Meter piping, lifting and mounting 27

and Backplane Board)

Page 42 of the Emerson 3812 user manual — BURST HAZARD
Page 42 · click to enlarge
Page 42

Section 2: Mechanical installation Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

2.2 Piping recommendations

BURST HAZARD

Before pipeline cleaning and maintenance (“pigging operations”), remove straightening vanes or flow conditioners.

Failure to do so may cause excessive pressure in the meter system, resulting in serious injury/ death or equipment damage.

3812 Ultrasonic Flow Meter with flow conditioner for unidirectional flow

Emerson 3812 — BURST HAZARD diagram (page 42)
Fig. 13

3812 Ultrasonic Flow Meter with flow conditioner for bidirectional flow

Emerson 3812 — BURST HAZARD diagram (page 42)
Fig. 14

28 Piping recommendations

Page 43 of the Emerson 3812 user manual — SUNSHIELD PROTECTION
Page 43 · click to enlarge
Page 43

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 2: Mechanical installation 3-9000-765 Rev H May 2015

Sunshields, provided by the customer, may be required to prevent exceeding the process fluid temperature when the meter is mounted in a location with extremely hot climates.

SUNSHIELD PROTECTION

Install a sunshield to prevent prolonged exposure to direct sunlight in extreme climates.

Failure to shield the meter may result in exceeding the process temperature range and damage transmitter electronics.

For optimal flow measurement conditions, Daniel suggests the piping configurations below. Regardless of the configuration selected, the user agrees to accept full responsibility for the site piping design and installation.

Flow conditioning is recommended for best measurement results.

  • Honed or un-honed meter tube(s)
  • Flow direction (unidirectional or bidirectional)
  • Correct meter size selection - too low may cause poor flow stability (thermal convection or too fast may cause erosion problems and resonance, cracks or failure of probes or thermowells (approximately .6 to 12 m/sec or 2 to 40 ft/sec).
  • Space availability for meter lengths (to allow inlet piping customization)
  • Concentric alignment pins or flange concentricity technique considerations

Figure 2-5 Piping recommendations unidirectional flow

Emerson 3812 — SUNSHIELD PROTECTION diagram (page 43)
Fig. 15

Piping recommendations 29

Page 44 of the Emerson 3812 user manual — 2.2 Piping recommendations
Page 44 · click to enlarge
Page 44

Section 2: Mechanical installation Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

Figure 2-6 Piping recommendations bidirectional flow

Emerson 3812 — 2.2 Piping recommendations diagram (page 44)
Fig. 16 · 2.2 Piping recommendations

All pipe lengths are minimum:

  • D = Nominal pipe size in inches (i.e. 6" pipe size; 10 D = 60 in)
  • P = Pressure measurement location
  • T = Temperature measurement location

Refer to the ultrasonic meter product data sheet for piping information. The Liquid Ultrasonic Flow Meter Datasheet may be downloaded from the Daniel website:

http://www2.emersonprocess.com/en-US/brands/daniel/Flow/ultrasonics/Pages/ultrasonic-3812.aspx

To access the product datasheet, from the Daniel products page (above link), select the Daniel Liquid Ultrasonic Flow Meter link, click the Documentation tab, expand the Data Sheets - Bulletins - Catalogs tab, then select the Data Sheet.

30 Piping recommendations

Page 45 of the Emerson 3812 user manual — FAULTY METER INSTALLATION
Page 45 · click to enlarge
Page 45

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 2: Mechanical installation 3-9000-765 Rev H May 2015

Meter tube dimensions with tube bundle or profiler plate for uni-directional and bi-directional flow, the minimum straight pipe length is as follows:

Table 2-2 Piping recommendation for uni-directional or bi-directional flow

UniDirectional Flow BiDirectional Flow

8D up stream (with a flow conditioner)

8D up stream (with a flow conditioner)

8D up stream (no flow conditioner)

8D up stream (no flow conditioner)

5D in front of flow conditioners if used 5D in front of flow conditioners if used

  • The bore of the mating piping should be within 1% of the meter inside diameter.
  • The meter is provided with dowel pins to align the meter body bore with the bore of the mating piping.
  • The Daniel Liquid Ultrasonic Flow Meter should be mounted in horizontal piping with the chord paths horizontal

FAULTY METER INSTALLATION

Correctly install the meter.

If meter bodies are mounted or oriented differently than specified above, debris may collect in the transducer ports which could adversely affect the transducer signals, or cause equipment damage.

  • Normally, the meter body is installed so that the electronics assembly is on the top of the meter. If there is insufficient space above the piping for this arrangement, the meter can be ordered with extra long transducer cables for remote mounting or the meter housing can be installed with the electronics assembly on the bottom.
  • The mating piping should include temperature and pressure measurement connections located a minimum of two nominal pipe diameters length down stream of the meter, or per API MPMS 5.8.

Piping recommendations 31

Page 46 of the Emerson 3812 user manual — CRUSHING HAZARD
Page 46 · click to enlarge
Page 46

Section 2: Mechanical installation Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

2.3 Meter safety for hoist rings and lifting slings

A Daniel Ultrasonic Meter can be safely lifted and maneuvered into and out of a meter run for installation or service by obeying the following instructions.

DANGER TO PERSONNEL AND EQUIPMENT

Lifting a Daniel Ultrasonic Meter with other equipment

The following lifting instructions are for installation and removal of the Daniel Ultrasonic Meter ONLY. The instructions below do not address lifting the Daniel ultrasonic meter while it is attached, bolted, or welded to meter tubes, piping, or other fittings.

Using these instructions to maneuver the Daniel Ultrasonic Meter while it is still attached, bolted, or welded to a meter tube, piping, or other fitting may result in equipment damage, serious injury, or death.

The operator must refer to their company's hoisting and rigging standards, or the "DOE-STD- 1090-2004 Hoisting and Rigging" standard if such company standards do not exist, for lifting and maneuvering any assembled meter tube and associated piping.

CRUSHING HAZARD

During meter installation or removal, always place the unit on a stable platform or surface that supports its assembled weight.

Failure to do so could allow the meter to roll, resulting in serious injury or equipment damage.

Prior to lifting the unit, refer to the Daniel 3812 Liquid Ultrasonic Flow Meter nameplate or outline dimensional (general arrangement) drawing for the assembled weight.

32 Meter safety for hoist rings and lifting slings

Page 47 of the Emerson 3812 user manual — 2.3.1 Use of appropriate safety engineered swivel hoist rings in meter end flanges
Page 47 · click to enlarge
Page 47

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 2: Mechanical installation 3-9000-765 Rev H May 2015

When lifting a Daniel Ultrasonic Meter by itself, Daniel recommends two methods. These methods are:

  • Using appropriately rated Safety Engineered Swivel Hoist Rings installed in the Daniel Ultrasonic Meter end flanges.
  • Using appropriately rated lifting slings positioned at designated areas of the Daniel Ultrasonic Meter.

Both methods must be used in conjunction with all appropriate company hoisting and rigging standards or the DOE-STD-1090-2004 HOISTING AND RIGGING standard if such company standards do not exist. Refer to the following sections for more information on these two methods.

2.3.1 Use of appropriate safety engineered swivel hoist rings in meter end flanges

Daniel Ultrasonic meters come equipped with a tapped hole located on the top of each meter body end flange. A flat machined surface surrounds each tapped hole (see Figure 2-7). This feature provides complete surface contact ONLY between the meter flange and an OSHA compliant Safety Engineered Swivel Hoist Ring as shown in Figure 2-8.

Operators SHALL NOT use Eye Bolts (see Figure 2-8) in the Daniel Ultrasonic Meter flange tapped holes to aid in lifting or maneuvering the unit.

Operators SHALL NOT use other Hoist Rings that do not fully seat flush with the counter bore on the top of the meter flanges.

Figure 2-7 Meter end flange with tapped flat-counterbore hole for hoist ring

Emerson 3812 — 2.3.1 Use of appropriate safety engineered swivel hoist rings in meter end flanges diagram (page 47)
Fig. 17 · 2.3.1 Use of appropriate safety engineered swivel ho…

A.

B.

A. Plug bolt

B. Flat counterbore surface

Use of appropriate safety engineered swivel hoist rings in meter end flanges 33

Page 48 of the Emerson 3812 user manual — Safety precautions using safety engineered swivel hoist rings
Page 48 · click to enlarge
Page 48

Section 2: Mechanical installation Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

Figure 2-8 Safety approved hoist ring and non-compliant eye bolt

Emerson 3812 — 2.3.1 Use of appropriate safety engineered swivel hoist rings in meter end flanges diagram (page 48)
Fig. 18 · 2.3.1 Use of appropriate safety engineered swivel ho…
Emerson 3812 — 2.3.1 Use of appropriate safety engineered swivel hoist rings in meter end flanges diagram (page 48)
Fig. 19 · 2.3.1 Use of appropriate safety engineered swivel ho…

Eye bolt Safety engineered swivel hoist ring

Safety precautions using safety engineered swivel hoist rings

Read and follow the Safety Precautions listed below:

1. Meters must only be lifted by personnel properly trained in the safe practices of rigging and lifting.
2. Remove the plug bolts installed in the tapped holes on the top of the flanges. Do not

discard the bolts as they must be reinstalled once the lifting operation is complete to prevent corrosion of the tapped holes.

3. Make sure the tapped holes on the meter are clean and free of debris before installing

the hoist rings.

4. Use only the safety engineered swivel hoist rings that are rated for lifting the meter. Do

not use any other type of hoist rings with the same screw size or heavy duty hoist rings. The meter tapping and counter bore size are suitable only for the hoist rings specified by Daniel.

5. When installing a hoist ring, make sure the base surface of the hoist ring fully contacts

the machined flat surface of the tapped hole. If the two surfaces do not come in contact then the hoist ring will not hold its full rated load. Torque the hoist ring attachment bolts to the limit indicated on the hoist rings.

6. After installation of the hoist rings, always check that the ring rotates and pivots freely

in all directions.

7. NEVER attempt to lift the meter using only one hoist ring.

34 Use of appropriate safety engineered swivel hoist rings in meter end flanges

Page 49 of the Emerson 3812 user manual — 2.3.1 Use of appropriate safety engineered swivel hoist rings in meter end flanges
Page 49 · click to enlarge
Page 49

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 2: Mechanical installation 3-9000-765 Rev H May 2015

8. Always use separate slings for each hoist ring. NEVER reeve one sling through both

hoist rings. The slings must be of equal length. Each sling must have a load rating that equals or exceeds the hoist ring load rating. The angle between the two slings going to the hoist rings must never exceed 90 degrees or the load rating of the hoist rings will be exceeded.

Figure 2-9 90 Degree angle between slings

Emerson 3812 — 2.3.1 Use of appropriate safety engineered swivel hoist rings in meter end flanges diagram (page 49)
Fig. 20 · 2.3.1 Use of appropriate safety engineered swivel ho…

Use of appropriate safety engineered swivel hoist rings in meter end flanges 35

Page 50 of the Emerson 3812 user manual — 2.3.1 Use of appropriate safety engineered swivel hoist rings in meter end flanges
Page 50 · click to enlarge
Page 50

Section 2: Mechanical installation Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

9. Direct mount option: NEVER allow the slings to contact the electronics enclosure.

Damage to the enclosure may occur. Use a spreader bar with the slings to prevent contact with the electronics enclosure and the base enclosure (see Figure 2-11). If the slings do come in contact with the electronic enclosure then remove the four bolts holding the enclosure to its base and temporarily remove the head from the meter during the lifting operation. You will need to unplug the cable on the Acquisition Module. Two screws hold this cable in place. Once the lifting operation is complete, reattach and secure the electronics cable on the Acquisition Module, return the electronics enclosure to its original position, replace the bolts, and secure the enclosure in place.

Lifting the meter with the upper enclosure installed but without the bolts installed, may cause the electronics to fall and cause personal injury or equipment damage.

Figure 2-10 Sling contacting electronics enclosure

Emerson 3812 — 2.3.1 Use of appropriate safety engineered swivel hoist rings in meter end flanges diagram (page 50)
Fig. 21 · 2.3.1 Use of appropriate safety engineered swivel ho…

36 Use of appropriate safety engineered swivel hoist rings in meter end flanges

Page 51 of the Emerson 3812 user manual — REMOVE CONDUIT TIE WRAPS FROM THE JUNCTION BOX
Page 51 · click to enlarge
Page 51

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 2: Mechanical installation 3-9000-765 Rev H May 2015

10. Remote mount option: Always use separate slings for each hoist ring. NEVER reeve one

sling through both hoist rings. The slings must be of equal length. Each sling must have a load rating that equals or exceeds the hoist ring load rating. The angle between the two slings going to the hoist rings must never exceed 90 degrees or the load rating of the hoist rings will be exceeded.

DO NOT DRAG THE TRANSMITTER ELECETRONICS ENCLOSURE DURING LIFTING OPERATIONS.

Support the transmitter electronics while lifting the meter body.

Failure to do so may r cause damage to the equipment.

REMOVE CONDUIT TIE WRAPS FROM THE JUNCTION BOX

Conduit tie wraps must be removed prior to powering the meter.

Tie wraps placed on the junction box conduit for protection during shipping must be removed before the meter is powered.

Use of appropriate safety engineered swivel hoist rings in meter end flanges 37

Page 52 of the Emerson 3812 user manual — 2.3.1 Use of appropriate safety engineered swivel hoist rings in meter end flanges
Page 52 · click to enlarge
Page 52

Section 2: Mechanical installation Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

11. NEVER apply shock loads to the meter. Always lift the meter gradually. If shock loading

ever occurs, the hoist ring must be inspected per manufacturer's recommendations prior to be placed in any further service. If a proper inspection cannot be performed, discard the hoist ring.

12. NEVER lift with any device, such as hooks, chains, or cables that could create side pulls

that could damage the ring of the hoist ring.

13. NEVER lift more than the ultrasonic meter assembly including electronics and

transducers with the hoist rings. The only exception is that it is safe to lift the meter with one ASME B16.5 or ASME B16.47 blind flange bolted to each end flange of the meter. NEVER use the hoist rings on the meter to lift other components such as meter tubes, piping or fittings attached to the meter. Doing so will exceed the load rating of the hoist rings.

14. Remove the hoist rings from the meter after lifting is completed and store them in an

appropriate case or container per their manufacturer's recommendation.

15. Apply heavy lubricant or anti-seize to the threads of the plug bolts and reinstall the plug

bolts to keep the tapped holes free of debris and to prevent corrosion.

38 Use of appropriate safety engineered swivel hoist rings in meter end flanges

Page 53 of the Emerson 3812 user manual — What size safety engineered swivel hoist ring do you need?
Page 53 · click to enlarge
Page 53

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 2: Mechanical installation 3-9000-765 Rev H May 2015

How to obtain safety engineered swivel hoist rings

A list of approved manufacturers of safety engineered hoist rings is below:

  • American Drill Bushing Company( <a href="http://www.americandrillbushing.com" rel="nofollow noopener" target="_blank">www.americandrillbushing.com</a> )
  • Carr Lane Manufacturing Company ( <a href="http://www.carrlane.com" rel="nofollow noopener" target="_blank">www.carrlane.com</a> )

Select an approved supplier from the list below. These vendors can supply the safety- engineered hoist rings. This is not intended to be a complete list.

  • Fastenal ( <a href="http://www.fastenal.com" rel="nofollow noopener" target="_blank">www.fastenal.com</a> )
  • Reid Tools (<a href="http://www.reidtool.com" rel="nofollow noopener" target="_blank">www.reidtool.com</a> )

The appropriate hoist rings can also be purchased directly from Daniel. The following table provides part numbers for reference:

Table 2-3 Hoist ring part number lookup table

Daniel part number1

Hoist ring thread size & load rating1

American Drill Bushing Co. P/N1

Carr Lane Manufacturing Co. P/N1

1-504-90-091 3/8"-16UNC, 1000 lb. 23053 CL-1000-SHR-1

1-504-90-092 1/2"-13UNC, 2500 lb 23301 CL-23301-SHR-1

1-504-90-093 3/4"-10UNC, 5000 lb. 23007 CL-5000-SHR-1

1-504-90-094 1"-8UNC, 10000 lb. 23105 CL-10000-SHR-1

1-504-90-095 1-1/2"-6UNC, 24000 lb. 23202 CL-24000-SHR-1

1. Note: The part numbers include only one hoist ring. Two hoist rings are required per meter.

What size safety engineered swivel hoist ring do you need?

To determine the size of the hoist rings required for your meter, use the appropriate table below for Liquid Ultrasonic Meters (see Table 2-3). Look down the column that matches the ANSI rating of your meter. Find the row that contains your meter size. Follow the row to the end to find the appropriate hoist ring part number.

Table 2-4 Hoist ring lookup table for Daniel Model 3812 Liquid Ultrasonic Flow Meters

ANSI 150 ANSI 300 Daniel Part Number

4”to 10” 4” to 10” 1-504-90-091

Use of appropriate safety engineered swivel hoist rings in meter end flanges 39

Page 54 of the Emerson 3812 user manual — Safety precautions using appropriate rated lifting slings
Page 54 · click to enlarge
Page 54

Section 2: Mechanical installation Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

2.3.2 Appropriately rated lifting slings

The following instructions are intended to provide general guidelines for proper lifting slings of the Daniel 3812 Ultrasonic meter by itself. These instructions are intended to be followed in addition to your company's standards or the DOE-STD-1090-2004 Hoisting and Rigging standard if such company standards do not exist.

Safety precautions using appropriate rated lifting slings

1. Meters must only be lifted by personnel properly trained in the safe practices of rigging and lifting.
2. NEVER attempt to lift the meter by wrapping slings around the electronics enclosure.

40 Appropriately rated lifting slings

Page 55 of the Emerson 3812 user manual — 2.3.2 Appropriately rated lifting slings
Page 55 · click to enlarge
Page 55

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 2: Mechanical installation 3-9000-765 Rev H May 2015

3. NEVER attempt to lift the meter using only one sling around the meter. Always use two

slings wrapped around each end of the body as shown below. A choker style sling is recommended using a spreader bar.

Figure 2-11 Correct sling attachment

Emerson 3812 — 2.3.2 Appropriately rated lifting slings diagram (page 55)
Fig. 22 · 2.3.2 Appropriately rated lifting slings
4. Visually inspect the slings prior to use for any signs of abrasion or other damage. Refer

to the sling manufacturer's procedures for proper inspection of the particular sling you are using.

5. Only use slings with ratings that exceed the weight to be lifted. Reference your

company's standards for safety factors that must be included when calculating the load rating.

Appropriately rated lifting slings 41

Page 56 of the Emerson 3812 user manual — 2.3.2 Appropriately rated lifting slings
Page 56 · click to enlarge
Page 56

Section 2: Mechanical installation Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

6. NEVER allow the slings to contact the electronics enclosure or the transducer shrouds.

Damage to the meter may occur. If the slings do come in contact with the electronics or the shroud, temporarily remove the head from the meter during the lifting operation (remove the four bolts holding the enclosure to its base and unplug the cable from the Acquisition Module. Two screws hold this cable in place.) Use a spreader-bar on the sling to prevent contact with the electronics or the transducer shroud.

7. Once the lifting operation is complete, reattach and secure the electronics cable to J3

on the Acquisition Module, return the electronics enclosure to its original position, replace the bolts, and secure the enclosure in place. Lifting the meter with the upper enclosure installed but with out the bolts installed, may cause the electronics to fall and cause personal injury or electronics damage.

Figure 2-12 Incorrect sling attachment

Emerson 3812 — 2.3.2 Appropriately rated lifting slings diagram (page 56)
Fig. 23 · 2.3.2 Appropriately rated lifting slings
8. NEVER apply shock loads to the meter. Always lift the meter gradually. If shock loading

ever occurs, the slings must be inspected per manufacturer's procedures prior to being placed in any further service.

42 Appropriately rated lifting slings

Page 57 of the Emerson 3812 user manual — 2.4 Mounting requirements in heated or cooled
Page 57 · click to enlarge
Page 57

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 2: Mechanical installation 3-9000-765 Rev H May 2015

2.4 Mounting requirements in heated or cooled

pipelines

The ambient operating temperature of the 3812 Liquid Allocation electronics (i.e. Flameproof enclosure and Intrinsically safe base enclosure) is -40 oC to +60 oC (-40 oF to +140 oF).

If the meter is installed into a pipeline which is heated or cooled outside this temperature range it is necessary to remove the Transmitter Electronics enclosure from the meter body (i.e. Spool piece acting as process fluid conduit) and mount it next to the meter body on a pipe stand or other rigid structure. The process temperature must also not exceed the operating temperature range of the transducers. LT-10 and LT-11 operating temperature range -50 oC to 135 oC (-58 oF to +275 oF) with NBR o-rings and operating temperature range -40 oC to +150 oC (-40 oF to +302 oF with FKM o-rings).

SURFACE TEMPERATURE HAZARD.

The meter body and piping may be extremely hot or cold.

Wear appropriate personal protective equipment when coming in contact with the meter.

Failure to do so may result in injury.

Mounting requirements in heated or cooled pipelines 43

Page 58 of the Emerson 3812 user manual — 2.4 Mounting requirements in heated or cooled pipelines
Page 58 · click to enlarge
Page 58

Section 2: Mechanical installation Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

44 Mounting requirements in heated or cooled pipelines

Page 59 of the Emerson 3812 user manual — Ω Ω Ω
Page 59 · click to enlarge
Page 59

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 3: Electrical installation 3-9000-765 Rev H May 2015

Section 3: Electrical installation

3.1 Cable length TTL mode

The maximum cable length is 2000 feet when the “TTL” mode is selected.

3.2 Cable length Open Collector mode

For the “open collector” mode, the maximum cable length depends on the cable parameters, pull-up resistance used, the maximum frequency to output, and frequency input parameters being driven. The following table provides estimated cable lengths for different pull-up resistor values and different Max Frequency settings in the meter using the following cable parameters. The table also provides an estimated cable voltage drop which indicates how much voltage will be across the cabling and effectively indicates to what voltage level the frequency input can be pulled down to by the frequency output.

If the voltage drop is higher than the voltage required for the frequency input to see a low state, then the configuration will most likely not work for your system. Performance of frequency outputs will vary from this table with setup and frequency input being driven.

Table 3-1 Configurations for open collector frequency outputs

Length (2 Conductors) Capacitance Resistance Resistance Frequency Current (2 Conductors)

(x1000ft) nF (Hz) (A) VDC

Ω Ω Ω

0.5 16.8 10.00 1000 1016.8 5000 0.024 0.397

1 33.6 20.00 1000 1033.6 1000 0.023 0.780

2 67.2 40.00 1000 1067.2 1000 0.022 1.511

4 134.4 80.00 1000 1134.4 1000 0.021 2.843

0.5 16.8 10.00 500 516.8 5000 0.046 0.780

1 33.6 20.00 500 533.6 5000 0.045 1.511

1.7 57.12 34.00 500 557.12 5000 0.043 2.461

6.5 218.4 130.00 500 718.4 1000 0.033 7.296

The 22 AWG wire characteristics are as follows:

  • Capacitance = 20 pF/ft or 20 nF/1000 ft (between two wires)
  • Resistance = 0.0168 Ohms/ft or 16.8 Ohms/1000 ft
  • Pull-up voltage is 24 VDC

Cable length TTL mode 45

Show 5 more labels / short text on this page
Cable resistance Cable · Pull-up resistance Total · Maximum frequency Sink · Cable voltage drop · Cable
Page 60 of the Emerson 3812 user manual — 3.3 Grounding meter electronics housing
Page 60 · click to enlarge
Page 60

Section 3: Electrical installation Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

3.3 Grounding meter electronics housing

Daniel 3812 Liquid Ultrasonic Flow Meter electronics should be internally grounded for intrinsically safe operations. Connect a wire to the chassis ground lug installed inside the Transmitter Electronics Enclosure as the primary ground. A secondary ground is located outside of the Transmitter Electronics Enclosure (see Figure 3-2). Digital grounds should never be connected to chassis ground.

The internal grounding terminal shall be used as the primary equipment ground. The external terminal is only a supplemental bonding connection where local authorities permit or require such a connection.

Figure 3-1 Transmitter Electronics Enclosure internal chassis ground

Emerson 3812 — 3.3 Grounding meter electronics housing diagram (page 60)
Fig. 24 · 3.3 Grounding meter electronics housing

A.

A. Transmitter Electronics Enclosure internal ground lug

46 Grounding meter electronics housing

Page 61 of the Emerson 3812 user manual — DEATH OR SERIOUS INJURY MAY OCCUR
Page 61 · click to enlarge
Page 61

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 3: Electrical installation 3-9000-765 Rev H May 2015

Figure 3-2 External ground lug

Emerson 3812 — 3.4 Conduit seals diagram (page 61)
Fig. 25 · 3.4 Conduit seals

A.

A. External ground lug

3.4 Conduit seals

Daniel 3812 Liquid Ultrasonic Flow Meters require conduit seals for installations in hazardous environments. Adhere to safety instructions to protect personnel and equipment.

HAZARDOUS VOLTAGE INSIDE

Do not open the Transmitter Electronics Enclosure when an explosive gas atmosphere is present. Disconnect equipment from supply circuit before opening.

Failure to remove power may result in serious injury or death.

DEATH OR SERIOUS INJURY MAY OCCUR

Explosion hazard. Substitution of components may impair intrinsic safety.

Do not disconnect equipment unless power has been removed or the area is known to be non-hazardous.

Conduit seals 47

Page 62 of the Emerson 3812 user manual — HAZARDOUS VOLTAGE INSIDE
Page 62 · click to enlarge
Page 62

Section 3: Electrical installation Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

3.4.1 Startup for systems using explosion-proof conduit

1. Assemble flexible conduit (recommended by Daniel) to the Transmitter Electronics Enclosure. A conduit seal fitting is required within 18 inches (457 mm) of the enclosure.
2. Check to make certain that all power to field wiring is turned OFF.

HAZARDOUS VOLTAGE INSIDE

Do not open the Transmitter Electronics Enclosure when an explosive gas atmosphere is present. Disconnect equipment from supply circuit before opening the enclosure.

Failure to remove power may result in serious injury or death.

3. Remove the end cap nearest the conduit entry to gain access to the transmitter

electronics.

4. Pull the wires.
5. Complete the field connection wiring.
6. Apply electrical power to the system and verify the field connections are working

correctly. Allow the system to run for the time specified by the customer (approximately one week) and an electrician has fully tested the connections. After the Acceptance Test is witnessed and approved, seal the conduit.

7. Power down the system and apply the sealing compound to the conduit and allow the

compound to set in accordance with manufacturer specifications.

8. Install the security latches and wire seals on the Transmitter Electronics Enclosure end

caps (see Section 3.7.1)

9. If required, install the wire seals on the Base Enclosure hex head bolts and on the Shroud

covering the meter body (see Section 3.7.2).

10. Install the security wire seals on the shrouds covering the transducers and cables. Refer

to the section appropriate for your meter design:

  • Section 3.7.4 “Bolted band shroud security seals”
  • Section 3.7.5 “Clamped band shroud security seals”
  • Section 3.7.6 “Split shroud security seals”
  • Section 3.7.7 “Latched band shroud security seals”
11. Re-apply electrical power to the system.
12. Set or configure the meter parameters using Daniel MeterLink. For additional

installation information refer to the system wiring diagram (see Appendix A), Daniel MeterLink Software for Gas and Liquid Ultrasonic Meters Quick Start Manual ( P/N 3-9000-763) and use Daniel MeterLink Field Setup Wizard to complete the configuration.

48 Startup for systems using explosion-proof conduit

Page 63 of the Emerson 3812 user manual — HAZARDOUS VOLTAGE INSIDE
Page 63 · click to enlarge
Page 63

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 3: Electrical installation 3-9000-765 Rev H May 2015

3.4.2 Startup for systems that use flame-proof cable

HAZARDOUS VOLTAGE INSIDE

Do not open the Transmitter Electronics Enclosure when an explosive gas atmosphere is present. Disconnect equipment from supply circuit before opening.

Failure to remove power may result in serious injury or death.

1. Check to make certain that all field wiring power is turned OFF.
2. Remove the end cap nearest the cable entries to gain access to the transmitter

electronics.

3. Install the cable and cable gland.
4. Complete the field connection wiring.
5. Connect a flow computer to the communications line on the Daniel3812 Liquid

Ultrasonic Flow Meter.

6. Apply electrical power to the system to ensure the field connections are working

correctly. Allow the system to run for the time specified by the customer (usually one week) and an electrician has fully tested the connections. After the Acceptance Test is witnessed and approved, seal the conduit.

7. Power down the system and apply the sealing compound to the conduit and allow to

set in accordance with manufacturer specifications.

8. Install the security latches and wire seals on the Transmitter Electronics Enclosure end

caps (see Section 3.7.1)

9. Install the wire seals on the Base Enclosure hex head bolts and on the Shroud covering

the meter body (see Section 3.7.2).

10. Install the security wire seals on the shrouds covering the transducers and cables. Refer

to the section appropriate for your meter design:

  • Section 3.7.4 “Bolted band shroud security seals”
  • Section 3.7.5 “Clamped band shroud security seals”
  • Section 3.7.6 “Split shroud security seals”
  • Section 3.7.7 “Latched band shroud security seals”
11. Connect electrical power to the system.
12. Set or configure the meter using Daniel MeterLink. For additional installation

information refer to the system wiring diagram (see Appendix A), Daniel MeterLink Software for Gas and Liquid Ultrasonic Meters Quick Start Manual (P/N 3-9000-763) and use Daniel MeterLink Field Setup Wizard to complete the meter configuration.

Startup for systems that use flame-proof cable 49

Page 64 of the Emerson 3812 user manual — 3.5 Wiring and I/O
Page 64 · click to enlarge
Page 64

Section 3: Electrical installation Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

3.5 Wiring and I/O

Daniel MeterLink uses the TCP/IP protocol to communicate with the 960-24™ MSTS electronics instead of Modbus ASCII or RTU. The TCP/IP protocol only works across either Ethernet, RS-485 full duplex (i.e., 4-wire), or RS-232. Daniel MeterLink can communicate with multiple meters if they are multi-dropped using 4-wire full duplex RS-485 mode. The meter electronics are HART capable and provide communication flexibility with Daniel 3812 Liquid Ultrasonic Flow Meters.

The HART® output provides communication with other field devices (e.g., 475/375 Field Communicator and AMS™ Device Manager software) and ultimately, communicates key diagnostic information through PlantWeb® architecture.

If not using Ethernet, a full duplex serial connection is necessary for Daniel MeterLink to communicate with a Daniel 3812 Liquid Ultrasonic Flow Meter.

The meter’s electronics auto-detects the protocol used and automatically switches between TCP/IP, Modbus ASCII, and Modbus RTU so it is not necessary to make any meter configuration changes to change the protocol.

50 Wiring and I/O

Page 65 of the Emerson 3812 user manual — 3.5.1 CPU Module labeling and LED indicators
Page 65 · click to enlarge
Page 65

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 3: Electrical installation 3-9000-765 Rev H May 2015

3.5.1 CPU Module labeling and LED indicators

The meter’s metrology mode and the status of the data transfer from the Acquisition Module to the CPU Module is indicated via light-emitting diode (LED) status indicators. The Write PROT. switch protects the meter’s configuration,

Figure 3-3 CPU Module labeling and LED indicators

A. B. C. D. E.

F. G.

Emerson 3812 — 3.5.1 CPU Module labeling and LED indicators diagram (page 65)
Fig. 26 · 3.5.1 CPU Module labeling and LED indicators

A. Acquisition/Measurement mode B. Power C. RX (RS-485/RS-232) - receiving data D. LED 4 - not used E. LED 5 - not used F. TX (RS-485/RS-232) - transmitting data

G. Link (Eth1 Link) - user Ethernet connection

CPU Module labeling and LED indicators 51

Page 66 of the Emerson 3812 user manual — Table 3-2 CPU Module labeling and LED functions
Page 66 · click to enlarge
Page 66

Section 3: Electrical installation Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

Table 3-2 CPU Module labeling and LED functions

CPU Module switch Function Switch position indicator or LED

  • WRITE PROT.
  • Write-protect mode - with switch in the ON position (default setting) protects configuration and firmware overwrites.
  • To write configuration changes or download firmware to the meter change the switch to the OFF position
  • Switch position
  • ON - (default setting) enables writeprotection of the configuration and firmware
  • OFF - enables writing configuration changes or downloading firmware
  • DHCP
  • Dynamic Host Protocol Server - enables you to communicate with a Daniel meter that is not connected to a network.
  • When the CPU Module switch is in the ON position, the meter is enabled to act as a DHCP server for a single DHCP client connected to the Ethernet port using a crossover cable. This should be used for peer to peer connections only.
  • When the connection is made, select to use the Meter Name in the meter instead of the Meter Directory Name in order to keep all log files and configurations separate from each meter.
  • Switch position
  • ON - the meter is enabled to act as a DHCP server for a single DHCP client
  • OFF - disables the DHCP server
  • PORT A
  • PORT A override - RS-232 serves as an override during meter commissioning to establish com munications and in the event the user cannot communicate with the meter due to an inadver tent communication configuration change. The override period is for two minutes
  • Supports: - auto-detected ASCII (Start bit 1, Data Bit 7, Parity Odd/Even, Stop Bit 1) - RTU (Start Bit 1, Data Bit 8, Parity none, Stop Bit 1). - Modbus protocols
  • RS-232 Baud rate=19,200
  • Modbus ID=32
  • Switch position
  • ON - enables RS-232 PORT A override
  • OFF - (default setting) disables RS-232 PORT A

MEAS System color indicates metrology mode • Acquisition mode

  • LED status
  • Red flashing LED
  • Solid red the Acquisition Module not communicating with the CPU Module
  • Green flashing LED
  • Measurement mode
  • PWR
  • 3.3V Power Indicator
  • Solid Green

LED 4 • Not used

LED 5 • Not used

RX • RX signal (Port A for RS485 or RS232 communication) receiving data

  • Flashing green (when receiving data)

TX • TX signal (Port A for RS485 or RS232 communication) transmitting data

  • Flashing green (when transmitting data)
  • LINK
  • ETH1Link user Ethernet connection
  • Solid green

52 CPU Module labeling and LED indicators

Page 67 of the Emerson 3812 user manual — RX-
Page 67 · click to enlarge
Page 67

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 3: Electrical installation 3-9000-765 Rev H May 2015

Ethernet communications

The Ethernet port IP address, subnet mask, and gateway address are software-configurable. In addition, a meter can be configured to act as a DHCP (Dynamic Host Configuration Protocol) server to assign an IP address to a PC or laptop running Daniel MeterLink. The DHCP server facility is not intended to act as a general purpose DHCP server for a wider network. To this end, no user control is provided over the class or range of IP addresses the unit provides. A standard twisted pair (Cat-5) cable should be used for Ethernet wiring.

It is strongly recommended that the meter be configured using an independent (off-network) single host. After configuration of the meter, the DHCP option must be turned off if used on a LAN/WAN.

RESTRICT ETHERNET AND SERIAL CONNECTIVITY USAGE

Failure to restrict Ethernet and communication access to the Daniel 3812 Liquid Ultrasonic Flow Meter can result in, among other things, unauthorized access, system corruption, and/ or data loss.

User is responsible for ensuring that physical access and Ethernet or electronic access to the Daniel 3812 Liquid Ultrasonic Flow Meter is appropriately controlled and any necessary security precautions, such as, establishing a firewall, setting password permissions and/or implementing security levels.

Use ethernet cable (Daniel P/N 3-3400-079) to connect the PC to the meter.

Table 3-3 Ethernet cable to PC communication

Emerson 3812 — Table 3-3 Ethernet cable to PC communication diagram (page 67)
Fig. 27 · Table 3-3 Ethernet cable to PC communication

White w/Orange Stripe TX+

Orange w/White Stripe TX -

White w/Green Stripe RX+

Green w/White Stripe RX -

RX-

RX+

TX-

TX+

A DIN 41612 48-pin connector is the interface from the CPU Module to the Field Connection Board (male end located on the back of the Field Connection Board).

CPU Module labeling and LED indicators 53

Ethernet communication · Wire color CPU

Page 68 of the Emerson 3812 user manual — Table 3-4 Serial Port A parameters
Page 68 · click to enlarge
Page 68

Section 3: Electrical installation Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

Serial connections

Use serial cable (Daniel P/N 3-2500-401) to connect to a PC running Daniel MeterLink. The cable is designed for RS-232 communications which is the serial Port A default configuration (see Appendix A field wiring diagram, Daniel Drawing DMC - 005558). The DB-9 end of the cable plugs directly into the PC running Daniel MeterLink. The three wires on the other end of the cable connect to the CPU Module. The RED wire goes to RX, the WHITE wire goes to TX, and the BLACK wire goes to COM for the RS-485/RS-232 (Table 3-4 for Port A wiring).

When Belden wire No. 9940 or equivalent is used, the maximum cable length for RS-232 communications at 9600 bps is 88.3 meters (250 ft.) and the maximum cable length for RS-485 communication at 57600 bps is 600 meters (1970 ft.).

Port A supports a special override mode which forces the port to use known communication values (19200 baud, address 32, RS-232). Note that the protocol is auto-detected. This mode is expected to be used during meter commissioning (to establish initial communication) and in the event that the user cannot communicate with the meter (possibly due to an inadvertent communication configuration change). Alternately, when using Daniel MeterLink™ with an Ethernet port, use Ethernet cable (Daniel P/N 3-3400-079) to connect the PC.

Table 3-4 Serial Port A parameters

Port/Communication Description Common features

  • Port A (Standard)
  • RS-232
  • RS-485 Half Duplex
  • RS-485 Full Duplex
  • Typically used for general communications with a flow computer, RTU (Modbus slave) and radios.
  • Special override mode to force port configura tion to known settings.
  • Supports RTS/CTS handshaking with software-configurable RTS on/off delay times.
  • Factory default is RS-232, Address 32, 19200 baud.
  • Communications via Daniel MeterLink using RS-232 or RS-485 Full Duplex
  • Software configurable Modbus Address (1-247)
  • Auto-detects TCP/IP and ASCII or RTU Protocol - ASCII Protocol: Start Bits = 1, Data Bits=7 1
- Parityodd or even 1, Stop Bits =1 1
- Baud Rates1200, 2400, 9600, 19200, 38400, 57600, 115000 bps - RTU Protocol:

Start Bits = 1, Data Bits=81

- Paritynone, Stop Bits = 1 1
- Baud Rates1200, 2400, 9600, 19200, 38400, 57600, 115000 bps
  • Ethernet
  • Preferred port for diagnostic communication via Daniel MeterLink
  • 10 Mbps/100 Mbps
  • Modbus TCP/IP
1. Denotes auto-detected protocols.

If not using Ethernet, a full duplex serial connection is necessary for Daniel MeterLink to communicate with a Daniel 3812 Liquid Ultrasonic Flow Meter.

54 CPU Module labeling and LED indicators

Page 69 of the Emerson 3812 user manual — RX
Page 69 · click to enlarge
Page 69

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 3: Electrical installation 3-9000-765 Rev H May 2015

Figure 3-4 PC to meter serial connection wiring

Emerson 3812 — 3.6 I/O connections diagram (page 69)
Fig. 28 · 3.6 I/O connections

RX

TX

COM

3.6 I/O connections

The meter provides the I/O connections on the CPU Module.

Figure 3-5 CPU Module I/O connections

A. B. C. D.

Emerson 3812 — RX diagram (page 69)
Fig. 29

A. Frequency/Digital Output 2 B. Frequency/Digital Output 3 C. Analog Output 2, 4-20mA output D. Analog Input - HART temperature and pressure connections

DB-9 Connector Female · I/O connections 55

Page 70 of the Emerson 3812 user manual — Frequency or Digital Outputs (FODO 2) source
Page 70 · click to enlarge
Page 70

Section 3: Electrical installation Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

3.6.1 Frequency/Digital outputs

The meter has three user-configurable selections for configuring either a Frequency output or Digital output (FODO).

  • FODO1 (four possible parameter configurations)
  • FODO2 (eight possible parameter configurations)
  • FODO3 (eight possible parameter configurations)

Frequency or Digital Outputs (FODO 1) source

  • FO1A, DO1A, FO1B, DO1B
  • Frequency output 1A is based on frequency content (Actual - Uncorrected Flow Rate)
  • Frequency output 1B is based on frequency content and Frequency 1 B Phase
  • Digital output 1A is based on Digital output1A content (Frequency1A Validity and Flow Direction)

Frequency or Digital Outputs (FODO 2) source

  • FO1A, DO1A, FO1B, DO1B, FO2A, DO2A, FO2B, DO2B
  • Frequency output 1A is based on frequency content (Actual - Uncorrected Flow Rate)
  • Frequency output 1B is based on frequency content and Frequency 1B Phase
  • Frequency output 2A is based on frequency content (Actual - Uncorrected Flow Rate)
  • Frequency output 2B is based on frequency content and Frequency 2B Phase
  • Digital output 1A is based on Digital output1A content (Frequency 1A Validity and Flow Direction)
  • Digital output 2A is based on Digital output2A content (Frequency 1A Validity and Flow Direction)
  • Digital output 2A is based on Digital output 2A content (Frequency 2A Validity and Flow Direction)
  • Digital output 2B is based on Digital output 2B content (Frequency 2B Validity and Flow Direction)

56 Frequency/Digital outputs

Page 71 of the Emerson 3812 user manual — Phase A and Phase B output (based on flow direction)
Page 71 · click to enlarge
Page 71

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 3: Electrical installation 3-9000-765 Rev H May 2015

Frequency or Digital Outputs (FODO 3) source

  • FO1A, DO1A, FO1B, DO1B, FO2A, DO2A, FO2B, DO2B
  • FO1A, DO1A, FO1B, DO1B, FO2A, DO2A, FO2B, DO2B
  • Frequency output 1A is based on frequency content (Actual - Uncorrected Flow Rate)
  • Frequency output 1B is based on frequency content and Frequency 1B Phase
  • Frequency output 2A is based on frequency content (Actual - Uncorrected Flow Rate)
  • Frequency output 2B is based on frequency content and Frequency 2B Phase
  • Digital output 1A is based on Digital output1A content (Frequency 1A Validity and Flow Direction)
  • Digital output 2A is based on Digital output2A content (Frequency 1A Validity and Flow Direction)
  • Digital output 2A is based on Digital output 2A content (Frequency 2A Validity and Flow Direction)
  • Digital output 2B is based on Digital output 2B content (Frequency 2B Validity and Flow Direction)

Mode options

  • Open Collector (requires external excitation supply voltage and pull-up resistor)
  • TTL (internally powered by the meter 0-5 VDC signal)

Channel B Phase options:

  • Lag forward, Lead reverse (Phase B lags Phase A while reporting forward flow, leads Phase A while reporting reverse flow)
  • Lead forward, Lag reverse (Phase B leads Phase A while reporting forward flow, lags Phase A while reporting reverse flow)

Phase A and Phase B output (based on flow direction)

  • Reverse flow - output only reports flow in the reverse direction. For frequency outputs, Phase B of the output is 90 degrees out of phase with Phase A.
  • Forward flow - output only reports flow in the forward direction. For frequency outputs, Phase B of the output is 90 degrees out of phase with Phase A.
  • Absolute - output reports flow in both directions. For frequency outputs, Phase B of the output is 90 degrees out of phase with Phase A.
  • Bidirectional - output reports flow on Phase A only in the forward direction and on Phase B only in the reverse direction.

Frequency/Digital outputs 57

Page 72 of the Emerson 3812 user manual — FO2A FO2B DO2A DO2B
Page 72 · click to enlarge
Page 72

Section 3: Electrical installation Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

Maximum frequency for the frequency outputs

  • 1000Hz
  • 5000Hz

Table 3-5 Frequency/Digital Outputs possible configurations

Frequency/Digital output Source configuration

Frequency /Digital Output 11

  • Frequency output 1A
  • Frequency output 1B
  • Digital output 1A
  • Digital output 1B

FODO1

FO1A FO1B DO1A DO2B

FODO2

Frequency /Digital Output 22

  • Frequency output 1A
  • Frequency output 1B
  • Digital output 1A
  • Digital output 1B
  • Frequency output 2A
  • Frequency output 2B
  • Digital output 2A
  • Digital output 2B

FO2A FO2B DO2A DO2B

FODO3

Frequency /Digital Output 32

1. Solid blue line denotes valid selection for Frequency/Digital Output 1. 2. Black dashed -line denotes valid selections for Frequency/Digital Output 2 and Frequency/Digital Output 3.

or · 58 Frequency/Digital outputs

Page 73 of the Emerson 3812 user manual — 3.6.1 Frequency/Digital outputs
Page 73 · click to enlarge
Page 73

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 3: Electrical installation 3-9000-765 Rev H May 2015

Output for FODO1 and Digital Output1 (Group 1 on the CPU Module) share a common ground and have 50V isolation. FODO2 and FODO3 (Group 2 on the CPU Module) share a common ground and have 50V isolation. This allows an output to be connected to a different flow computer. The outputs are opto-isolated from the CPU Module and have a withstand voltage of at least 500V rms dielectric.

Figure 3-6 CPU Module - Frequency/Digital outputs common ground

Emerson 3812 — 3.6.1 Frequency/Digital outputs diagram (page 73)
Fig. 30 · 3.6.1 Frequency/Digital outputs

A.

B.

A. FODO1 and Digital input1 - shared common ground (Group 1) B. FODO2 and FODO3 - shared common ground (Group 2)

Frequency/Digital outputs 59

Page 74 of the Emerson 3812 user manual — DHCP OFF ON
Page 74 · click to enlarge
Page 74

Section 3: Electrical installation Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

3.6.2 Analog input settings

The 960-24™ MSTS has the capability to sample analog temperature (Analog Input 1) and pressure (Analog Input 2) with 4-20 mA signals. These analog input signals are configured to sink. The two independent analog input circuits are configured for conventional 4-20 mA service. Also, 24VDC isolated power supply connection is provided for an external power source. Refer to the Field wiring diagram (see Appendix A drawing DMC004936).

3.6.3 Analog output settings

The 960-24™ MSTS provides two 4-20 mA analog output signals that are software configurable for either sink or source current (see Appendix A drawing DMC004936).

Full HART® functionality is provided so that any commercially available HART® transmitter which meets the specifications of the HART® Communications Foundation can be connected to the Daniel 3812 Liquid Ultrasonic Flow Meter.

Analog Output 2 (AO2) is user-configurable as a conventional 4-20 mA output.

3.6.4 Digital Input

The Daniel 3812 Liquid Ultrasonic Flow Meter provides one digital input that can be used as a general purpose input or used for synchronizing calibration (for flow calibration gating - contact closure). The meter records the volume seen between switch closures. The polarity of the input is configured as normal or inverted polarity.

  • polarity is determined by the IsDI1ForCalActiveLow and the gating edge is determined by the IsDI1ForCalStateGated (calibrate edge gated or calibrate state gated).
  • calibration is started and stopped via an inactive>active state change.

The digital input must be configured via the Daniel MeterLink Tools>Edit/Compare Configuration page.

3.6.5 DHCP server switch settings

The meter can be configured to act as a DHCP server. The DHCP server is enabled/disabled via CPU Module DHCP switch as follows:

Table 3-6 DHCP server switch settings

CPU Module switch DHCP server disabled DHCP server enabled

DHCP OFF ON

60 Analog input settings

Page 75 of the Emerson 3812 user manual — 3.6.7 External power source connection and fuse
Page 75 · click to enlarge
Page 75

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 3: Electrical installation 3-9000-765 Rev H May 2015

3.6.6 Configuration protect switch settings

The meter’s configuration parameters and firmware can be protected against changes via CPU Module Write PROT. switch as follows:

Table 3-7 Configuration protect switch settings

CPU Module switch Configuration protected Configuration unprotected

WRITE PROT. ON (default) OFF

A complete list of write-protected parameters are in Appendix A.

3.6.7 External power source connection and fuse

Located inside the Transmitter Electronics Enclosure is a connector for a user-provided external power source, a 2 Ampere fuse and a 24V loop power connection for ultrasonic meter analog outputs, generator transmitter or pressure transmitter devices. The current is limited to 88mA.

Figure 3-7 CPU Module power source connections

A. B.

Emerson 3812 — 3.6.7 External power source connection and fuse diagram (page 75)
Fig. 31 · 3.6.7 External power source connection and fuse

A. Power In connector (main power) B. 24V LOOP POWER C. 2 Ampere fuse (used for the main power input)

Configuration protect switch settings 61

Page 76 of the Emerson 3812 user manual — 3.7.1 Direct or remote mount transmitter electronics enclosure seal
Page 76 · click to enlarge
Page 76

Section 3: Electrical installation Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

3.7 Security seal installation

Security seals protect the integrity of the meter metrology and prevent tampering with transducer assemblies. The following sections detail how to properly seal the Daniel 3812 Liquid Ultrasonic Flow Meter after commissioning. The security seal wires are commercially available.

CUTTING HAZARD

Sharp edges may be present on the band shrouds.

Wear appropriate personal protective equipment when working on the meter. Failure to do so may cause serious injury.

Be sure to set the WRITE PROT. switch on the CPU Module to the ON position prior to sealing the enclosure.

3.7.1 Direct or remote mount transmitter electronics enclosure seal

Use the following instructions to install the security seal wires on the Transmitter Electronics Enclosure.

Figure 3-8 Transmitter electronics enclosure security latch

Emerson 3812 — 3.7.1 Direct or remote mount transmitter electronics enclosure seal diagram (page 76)
Fig. 32 · 3.7.1 Direct or remote mount transmitter electronics…

A.

B.

A. Transmitter Electronics Enclosure end cap B. Security latch

62 Security seal installation

Page 77 of the Emerson 3812 user manual — Procedure
Page 77 · click to enlarge
Page 77

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 3: Electrical installation 3-9000-765 Rev H May 2015

Procedure

1. Rotate the end cap clockwise fully closing and compressing the end cap seal. Install the Security latch using a 3mm Allen wrench.
2. Install the security seal wire into and through one of the two holes in the end cap.

Choose holes that minimize counterclockwise rotation of the end cap when the security wire is taut (maximum wire diameter .078 inch; 2.0mm).

Figure 3-9 Direct or remote mount Transmitter Electronics Enclosure security seals

Emerson 3812 — Procedure diagram (page 77)
Fig. 33 · Procedure

A.

B.

A. Transmitter Electronics Enclosure end cap B. Security wire seals

3. Adjust the security wire, removing all slack and thread into the lead seal.
4. Cut wire ends to remove excess wire.
5. This completes the direct or remote Transmitter Electronics Enclosure seal installation

procedure.

Direct or remote mount transmitter electronics enclosure seal 63

Page 78 of the Emerson 3812 user manual — Procedure
Page 78 · click to enlarge
Page 78

Section 3: Electrical installation Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

3.7.2 Base Enclosure Security Seals

Use the following instructions to install the security seal wire on the Base Enclosure.

Procedure

1. Install security wire seal into and through two of the four the holes in the socket head screws on the Base Enclosure cover (maximum wire diameter .078 inch; 2.0mm).

Figure 3-10 Base Enclosure security seals

Emerson 3812 — Procedure diagram (page 78)
Fig. 34 · Procedure

A.

B.

A. Base Enclosure B. Security wire seals

2. Position the wire to prevent counterclockwise rotation of the screws when the seal wire

is taut.

3. Twist and adjust wire removing all slack and seal.
4. Cut wire ends to remove excess wire.
5. This completes Base Enclosure security seal installation procedure.

64 Base Enclosure Security Seals

Page 79 of the Emerson 3812 user manual — Procedure
Page 79 · click to enlarge
Page 79

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 3: Electrical installation 3-9000-765 Rev H May 2015

3.7.3 Seal the Junction Box remote mount electronics option

Use the following instructions to install the security seal wire on the junction box, if required.

Procedure

1. Install security wire seal into and through two of the four the holes in the socket head screws on the Junction Box cover (maximum wire diameter .078 inch; 2.0 mm).

Figure 3-11 3812 Remote mount transmitter electronics option

Emerson 3812 — Procedure diagram (page 79)
Fig. 35 · Procedure

A.

A. Junction Box socket head screws

2. Position the wire to prevent counterclockwise rotation of the screws when the seal wire

is taut.

3. Twist and adjust wire removing all slack and seal.
4. Cut wire ends to remove excess wire.
5. This completes installing security seals on the Junction Box.

Seal the Junction Box remote mount electronics option 65

Page 80 of the Emerson 3812 user manual — 3.7.4 Bolted band shroud security seals
Page 80 · click to enlarge
Page 80

Section 3: Electrical installation Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

3.7.4 Bolted band shroud security seals

If required, use the following instructions to install bolted band shrouds security wire seals.

Figure 3-12 Bolted band shroud security seals - bottom view

Emerson 3812 — 3.7.4 Bolted band shroud security seals diagram (page 80)
Fig. 36 · 3.7.4 Bolted band shroud security seals

A.

B.

A. Band shroud

D. E.

B. Left front shroud bolt 1/4”- 20

C. Left back shroud bolt 1/4”- 20 D. Right back shroud bolt 1/4”- 20 E. Right front shroud bolt 1/4”- 20

66 Bolted band shroud security seals

Page 81 of the Emerson 3812 user manual — Procedure
Page 81 · click to enlarge
Page 81

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 3: Electrical installation 3-9000-765 Rev H May 2015

Procedure

1. Beneath the meter, install the security wires into and through the left front bolt (Item B) and feed through the holes in the right back bolt (Item D).
2. Repeat wire installation through the right front bolt (Item E) to the left rear bolt (Item C)

as shown in Figure 3-12. The maximum wire diameter .078 inch; 2.0 mm.

3. Position the wire to prevent counterclockwise rotation of the bolts when the seal wire is

taut.

4. Adjust the security wire, removing all slack and thread into the lead seal.
5. Cut wire ends to remove excess wire.
6. This completes the bolted band shroud security seal installation procedure.

3.7.5 Clamped band shroud security seals

Use the following instructions to install the security seal wires, if required, on the two top end shrouds covering the worm screw clamps. This procedure applies to clamped band shroud meters.

Figure 3-13 Clamped shroud security seals

Emerson 3812 — 3.7.5 Clamped band shroud security seals diagram (page 81)
Fig. 37 · 3.7.5 Clamped band shroud security seals

A.

B.

A. Top end shroud B. Security wire seals

Procedure

1. Install the security seal wire into and through the two bolt holes on the top end shrouds (maximum wire diameter .078 inch; 2.0 mm). Position the wire to prevent counter clockwise rotation of the screws when the seal wire is taut.
2. Remove all slack and seal.
3. Repeat previous step for the other top end shroud.
4. Cut wire ends to remove excess wire.
5. This completes the clamped band shroud security seal installation procedure.

Clamped band shroud security seals 67

Page 82 of the Emerson 3812 user manual — Procedure
Page 82 · click to enlarge
Page 82

Section 3: Electrical installation Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

3.7.6 Split shroud security seals

Use the following instructions to install the security seal wires on the split shroud covering the meter body and transducer assemblies.

Figure 3-14 Split shroud security seals

Emerson 3812 — 3.7.6 Split shroud security seals diagram (page 82)
Fig. 38 · 3.7.6 Split shroud security seals

A.

B

D

A. Upper split shroud B. Split shroud clamp

C. Security wire seals D. Lower split shroud

Procedure

1. Install the security seal wire into and through the holes in the split shroud clamp bent tab and the upper split shroud bent tab. (maximum wire diameter .078 inch; 2.0 mm).
2. Remove all slack and seal.
3. Cut wire ends to remove excess wire.
4. Repeat these steps for the other split shroud clamps.
5. This completes the split shroud security seal installation procedure.

68 Split shroud security seals

Page 83 of the Emerson 3812 user manual — 3.7.7 Latched band shroud security seals
Page 83 · click to enlarge
Page 83

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 3: Electrical installation 3-9000-765 Rev H May 2015

3.7.7 Latched band shroud security seals

Use the following instructions to install the security seal wires on the latched band shrouds covering the meter body and transducer assemblies.

Figure 3-15 Latched band shroud assembly

Emerson 3812 — 3.7.7 Latched band shroud security seals diagram (page 83)
Fig. 39 · 3.7.7 Latched band shroud security seals

A.

B.

D.

E.

F.

A. Band shroud B. Stainless steel strike

C. Transducer cable D. Meter body recess for pop rivet allowance

E. Meter body shoulder F. Shroud latch

Latched band shroud security seals 69

Page 84 of the Emerson 3812 user manual — Procedure
Page 84 · click to enlarge
Page 84

Section 3: Electrical installation Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

Procedure

1. Install the security seal wire into and through the holes in one of the shroud latches (maximum wire diameter .078 inch; 2.0 mm).

Figure 3-16 Shroud latch holes for security wire seals

Emerson 3812 — Procedure diagram (page 84)
Fig. 40 · Procedure

A.

B.

A. Left shroud latch holes for security wire seals

B. Right shroud holes for security wires

2. Remove all slack and seal.
3. Cut wire ends to remove excess wire.
4. Repeat these steps for the other shroud latch.
5. This completes the split shroud security seal installation procedure.

70 Latched band shroud security seals

Page 85 of the Emerson 3812 user manual — 3.7.8 Seal conduit ports
Page 85 · click to enlarge
Page 85

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 3: Electrical installation 3-9000-765 Rev H May 2015

3.7.8 Seal conduit ports

The unit should be properly sealed with a sealing compound after electrical connections have been tested according to the customer's Best Practices schedule. Some areas require a witnessed Acceptance Test for the installed system and require that the meter run for a predetermined length of time (approximately one to two weeks) before the unit is sealed. This allows time to verify all electrical connections are correct, that the meter is accurately measuring flow and that the meter meets the customer’s installation requirements. See Section 3.4.1 and Section 3.4.2.

Seal conduit ports 71

Page 86 of the Emerson 3812 user manual — 3.7.8 Seal conduit ports
Page 86 · click to enlarge
Page 86

Section 3: Electrical installation Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

72 Seal conduit ports

Page 87 of the Emerson 3812 user manual — 4.1 Daniel MeterLink setup
Page 87 · click to enlarge
Page 87

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 4: Configuration 3-9000-765 Rev H May 2015

Section 4: Configuration

After the mechanical and electrical installation is complete and connectivity is established, use the Daniel MeterLink Software for Gas and Liquid Ultrasonic Meters Quick Start Manual (P/N 3-9000-763) to setup initial communications with the meter.

4.1 Daniel MeterLink setup

1. Review the software operating system, hardware and peripheral requirements.
2. Follow the installation instructions for your operating system (Windows® XP, Windows

Vista, Windows®7 or Windows® 8).

3. Configure a direct connection driver for first time modem configuration for Daniel

MeterLink communications.

4. Select the Daniel MeterLink desktop icon and complete the information in the

Registration Wizard. This wizard will appear the when you first run Daniel MeterLink and will provide you the following options.

Options:

Register by e-mail • Select this option and click Next to register by e-mail. Your machine must be connected to the Internet and have a default mail utility (i.e. Microsoft® Office Outlook®, Outlook® Express, etc.) configured to use this option.

Register by phone • Select this option to register by phone. After clicking Next, a page will display with phone numbers to dial.

Register later (remind me) • Select this option if you do not wish to register now, but would still like the wizard to appear the next time you run MeterLink.

Register later using the Help | Register Program menu item (don’t remind me) • Select this option if you do not wish to register now and do not want this wizard to appear the next time you run MeterLink. You can still register at a later time by selecting Register Program from the Help menu

5. Select File>Program Settings and customize the user-preferences (e.g. User name,

Company name, display units, Liquid Meter volume units and other interface settings)

6. Connect to your meter. If your meter is not shown in the list, select Edit Meter Directory

and setup the connections properties.

7. Run the Field Setup Wizard.

Daniel MeterLink setup 73

Page 88 of the Emerson 3812 user manual — 4.2 Field Setup Wizard
Page 88 · click to enlarge
Page 88
Section 4Configuration Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

4.2 Field Setup Wizard

1. Use the Field Setup Wizard-Startup Page and select the checkboxes that allow proper configuration for your meter (Temperature, Pressure, Meter Corrections, and Meter Outputs). Selections on this page will affect other configuration selections. Select Next to continue to General setup.
2. On General setup page configure the meter’s system units (U.S Customary or Metric

units) volume units, flow rate time, low flow cutoff, contract hour and enable reverse flow. Click Next to continue to Frequency Outputs.

3. Configure Frequency output 1 and Frequency output 2 content (Daniel Liquid

Ultrasonic Meters content is Uncorrected flow rate), flow direction, Channel B phase, maximum frequency output (Hertz) and Full scale volumetric flow rate. Click Next to continue to Meter Digital Outputs.

4. Select the Meter Digital Output parameters for Digital output 1A, Digital output 1B,

Digital output 2A and Digital output 2B based on Frequency validity or flow direction. if the output of the ultrasonic meter is reversed from what a flow computer is expecting, select Inverted Operation. This changes the digital output from a HIGH for a TRUE condition to output a LOW for a TRUE condition. Click Next to continue to Current Outputs.

5. Current Outputs are based on Uncorrected (Actual) flow rate) content, flow direction

(Forward, Reverse or Absolute) and Full scale volumetric flow rate used with output (20mA maximum). Alarm action parameters determines the state the output will drive during an alarm condition (High 20mA, Low - 4 mA, Hold last value, Very low - 3.5, Very high 20.5 mA or None). Click Next to continue to configure the HART® Output(s) parameters.

6. HART® Output parameters include four Dynamic process variables (Primary,

Secondary, Third and Fourth variable. The Primary variable is set to match the Content set for Current output 1. If a second current output is available, the Secondary variable is set to match the Content set for Current output 1) Identification and HART® units (volume units, Flow rate time units, Velocity units, Pressure and Temperature units). Click Next to continue to Temperature and Pressure.

7. Set the temperature and pressure scaling for analog inputs, enter fixed values, and set

alarm limits for both.

8. Select Finish to write the configuration settings to the meter.
9. Configure the parameters for the local display. Use the drop-down arrow in the Display

Items list box and select or modify the Display items, the Display units and the Scroll delay.

74 Field Setup Wizard

Page 89 of the Emerson 3812 user manual — Table 4-1 Local display labels, descriptions and valid units
Page 89 · click to enlarge
Page 89

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 4: Configuration 3-9000-765 Rev H May 2015

Display Items

The valid labels, descriptions and units for the shown below:

Table 4-1 Local display labels, descriptions and valid units

Local Display labels, descriptions and units

QFLOW — Uncorrected volume flow rate

  • BBL – Barrels
  • GAL – Gallons
  • L – Liters
  • CM – Cubic Meters
  • MCM – Thousand Cubic Meters

TDYVL — Current day’s forward uncorrected volume

  • +BBL – Barrels
  • +GAL – Gallons
  • +L – Liters
  • +CM – Cubic Meters
  • +MCM – Thousand Cubic Meters

TDYVL — Current day’s reverse uncorrected volume

  • -BBL – Barrels
  • -GAL – Gallons
  • -L – Liters
  • -CM – Cubic Meters
  • -MCM – Thousand Cubic Meters

YSTVL — Previous day’s forward uncorrected volume

  • +BBL – Barrels
  • +GAL – Gallons
  • +L –Liters
  • +CM – Cubic Meters
  • +MCM – Thousand Cubic Meters

YSTVL — Previous day’s reverse uncorrected volume

  • -BBL – Barrels
  • -GAL – Gallons
  • -L – Liters
  • -CM – Cubic Meters
  • -MCM – Thousand Cubic Meters

TOTVL — Forward uncorrected volume

  • +BBL – Barrels
  • +GAL – Gallons
  • +L – Liters
  • +CM – Cubic Meters
  • +MCM – Thousand Cubic Meters

TOTVL — Reverse uncorrected volume

  • -BBL – Barrels
  • -GAL – Gallons
  • -L – Liters
  • -CM – Cubic Meters
  • -MCM – Thousand Cubic Meters

Field Setup Wizard 75

Page 90 of the Emerson 3812 user manual — Table 4-1 Local display labels, descriptions and valid units
Page 90 · click to enlarge
Page 90
Section 4Configuration Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

Table 4-1 Local display labels, descriptions and valid units

Local Display labels, descriptions and units

VEL — Average flow velocity

  • Ft/S – Feet per second
  • M/S – Meters per second

SOS — Average sound velocity

  • Ft/S – Feet per second
  • M/S – Meters per second

TEMP — Flow-condition temperature

  • DEGF – Degrees Fahrenheit
  • DEGC – Degrees Celsius

PRESS — Flow-condition pressure

  • PSI – Pound per square inch
  • MPA – Megapascals

FRQ1A — Frequency channel 1A

  • HZ – Hertz

FRQ1B — Frequency channel 1B

  • HZ – Hertz

KFCT1 — Frequency 1 K-factor

  • BBL – Barrels
  • GAL – Gallons
  • L – Liters
  • CM – Cubic Meters
  • MCM – Thousand Cubic Meters

FRQ2A — Frequency channel 2A

  • HZ – Hertz

FRQ2B — Frequency channel 2B

  • HZ – Hertz

KFCT2 — Frequency 2 K-factor

  • BBL – Barrels
  • GAL – Gallons
  • L – Liters
  • CM – Cubic Meters
  • MCM – Thousand Cubic Meters

AO1 — Analog Output 1 current

  • MA – Milliamperes

AO2 — Analog Output 2 current

  • MA – Milliamperes

76 Field Setup Wizard

Page 91 of the Emerson 3812 user manual — Scroll delay
Page 91 · click to enlarge
Page 91

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 4: Configuration 3-9000-765 Rev H May 2015

Display units

The Meter volume units displayed are either U.S. Customary or Metric. To modify the Display Units, configure the Meter units system in the Field Setup Wizard — General Page.

  • U.S. Customary volume unit selections are:
  • Metric volume unit selections are:
  • Display units preceded by a plus or minus sign indicate forward and reverse flow direction.
  • The Local Display Flow rate time units are modifiable by selecting the drop-down arrow and clicking the time unit in the list box.
  • Valid flow rate time units selections are:

Scroll delay

The Scroll Delay is the time interval for the selected display items to be shown on the Local Display. The default scroll delay setting is five seconds. Click the spin box up or down arrow to increase or decrease the length of time an item displays.

1. Select Finish to write the configuration settings to the meter.
2. Save the meter configuration file, collect a Maintenance log and Waveforms to

document the “As Left” settings.

Field Setup Wizard 77

Show 8 more labels / short text on this page
– Barrels · – Gallons · – Cubic meters · – Cubic liters · – second · – minute · – hour · – day
Page 92 of the Emerson 3812 user manual — 4.3 Using AMS Device Manager to configure the
Page 92 · click to enlarge
Page 92
Section 4Configuration Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

4.3 Using AMS Device Manager to configure the

meter

This procedure assumes you have AMS Device Manager installed on the host computer and have downloaded the latest Daniel Liquid Ultrasonic Meter Device Description (DD).

If not installed, click the link below to download the AMS device installation tool kit.

http://www2.emersonprocess.com/en-US/documentation/deviceinstallkits/Pages/devicein stallkitsearch.aspx

Procedure - installing AMS Device Description

1. Use the link above to search for the Device Description (DD) for your Daniel 3810 Series Liquid Ultrasonic Flow Meter.
2. Use the pull-down menu and select the Brand/Manufacturer - Emerson Daniel Industries .
3. Next select the Device, Liquid 3810 Series from the pull-down menu.
4. Choose the Device Revision1 , from the pull-down menu.
5. Next, select HART from the Communication Protocol menu.
6. Select AMS Device Manager for the Host System.
7. Select the Host System Revision 11.5 .
8. Verify your search parameters are correct, as shown below.

Figure 4-1 AMS Device Description search

Emerson 3812 — 4.3 Using AMS Device Manager to configure the diagram (page 92)
Fig. 41 · 4.3 Using AMS Device Manager to configure the
9. Click Search Now .

78 Using AMS Device Manager to configure the meter

Page 93 of the Emerson 3812 user manual — 4.3 Using AMS Device Manager to configure the meter
Page 93 · click to enlarge
Page 93

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 4: Configuration 3-9000-765 Rev H May 2015

10. Click the Daniel Industries Liquid 3810 Series Rev 1 hyperlink. The file download dialog

displays. Click the Save button to save the files to your host system. You may use the default download location or change the directory.

11. AMS file download options
Emerson 3812 — 4.3 Using AMS Device Manager to configure the meter diagram (page 93)
Fig. 42 · 4.3 Using AMS Device Manager to configure the meter
12. Click the Save button to complete the file download.

Figure 4-2 AMS file download complete

Emerson 3812 — 4.3 Using AMS Device Manager to configure the meter diagram (page 93)
Fig. 43 · 4.3 Using AMS Device Manager to configure the meter
13. Click Open or Open Folder to view the downloaded files.
14. Establish power to the meter and wiring to Analog Input 1 for HART communication.
15. Start the AMS Device Manager using a laptop or PC.
16. Enter login credentials and click OK to launch the application.

Using AMS Device Manager to configure the meter 79

Page 94 of the Emerson 3812 user manual — 4.3 Using AMS Device Manager to configure the meter
Page 94 · click to enlarge
Page 94
Section 4Configuration Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H
17. Click the Configure tab, and then select Guided Setup , Manual Setup or Alert Setup .

Figure 4-3 AMS Device Manager

Emerson 3812 — 4.3 Using AMS Device Manager to configure the meter diagram (page 94)
Fig. 44 · 4.3 Using AMS Device Manager to configure the meter

Figure 4-4 AMS Device Manager - Overview

Emerson 3812 — 4.3 Using AMS Device Manager to configure the meter diagram (page 94)
Fig. 45 · 4.3 Using AMS Device Manager to configure the meter

80 Using AMS Device Manager to configure the meter

Page 95 of the Emerson 3812 user manual — 4.3 Using AMS Device Manager to configure the meter
Page 95 · click to enlarge
Page 95

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 4: Configuration 3-9000-765 Rev H May 2015

AMS Device Manager - Guided Setup

The Guided setup wizard provides configuration parameter settings for the meter. The Guided Setup is a subset of the Manual Setup parameters.

Figure 4-5 AMS Device Manager - Guided Setup

Emerson 3812 — 4.3 Using AMS Device Manager to configure the meter diagram (page 95)
Fig. 46 · 4.3 Using AMS Device Manager to configure the meter
1. Disable the Write Protect switch in the CPU Module to write any of the following configuration parameters to your meter.
2. Click the Setup Units tab to configure the system units (U.S. Customary or Metric units),

Volume units, Flow rate time units, Velocity units, Pressure units and Temperature units. Click Apply to write the parameters to the meter.

3. Click the Setup Outputs tab to configure the Device Variables Mapping, Units,

Frequency/Digital outputs, Frequency and Digital Outputs 1 and 2, Analog outputs, Digital Input, Pressure and Temperature.

a. Analog output 1 (HART) - Content (Primary Variable) displays Uncorrected Flow Rate and is a read only attribute). Configure Direction (flow), Lower Range value, Upper range value and Alarm Action and view the HART Parameters Tag, Date, Descriptor, Message, Final Assembly Number Poll Address, Number of Response Preambles. b. Analog Output 2 - Content (Secondary Variable) displays Uncorrected Flow Rate and has a read only attribute. Configure Direction (flow), Lower Range value, Upper range value and Alarm Action. Map the Third and Four variables using the Manual Setup wizard. Selections include Uncorrected Volume Flow Rate, Pressure and Temperature.

Using AMS Device Manager to configure the meter 81

Procedure

Page 96 of the Emerson 3812 user manual — 4.3 Using AMS Device Manager to configure the meter
Page 96 · click to enlarge
Page 96
Section 4Configuration Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H
4. After all of the data shown below is entered, click Apply to write the parameters to the

meter.

a. Click the Frequency/Digital Outputs tab to configure Frequency/Digital Output 1, 2 and 3 Source and drive Mode. Select the Source for each Frequency/Digital output and select the desired drive Mode. The Mode options are Open Collector which requires an external excitation voltage and pull-up resistor or TTL mode which outputs a 0-5 VDC signal (each Frequency output has an A and B output phase). (Refresh Note: If changes are made to any Source variable on this page, apply the changes and navigate to the Guided Setup page. Navigate back to the Manual Setup for the changes to be reflected in other Manual Setup pages). b. Click the Frequency and Digital Output 1 tab to configure the Content, (flow) Direction, Channel B Phase frequency output, Lag forward, Lead Reverse or Lead Forward, Lag Reverse (Phase B lags Phase A while reporting forward flow and lead Phase A while reporting reverse flow or the opposite), Digital Output 1 Channel A Content and Polarity, Channel B Content and Polarity, Maximum Frequency, and Lower and Upper Range Units of Measure. c. Click the Frequency and Digital Output 2 tab and repeat Step 3b to configure Frequency and Digital Output 2 parameters.

5. Click Setup HART to configure the HART parameters (tag, date, descriptor, message

text, Final Assembly number, Poll address and number of response preambles are displayed). After all of the data is entered click Apply to write the parameters to the meter.

6. On the Overview page, click Alert Setup and select the Flow Analysis tab and enable

Reverse Flow. Click the OK button to return to the Overview page.

7. On the Overview page, click the Service Tools tab and select the Variables tab. The Flow

Data, Path Information, Flow Totals, and All Variables data is populated after you are connected to the meter.

a. Click the Flow Data tab and view the Flow Direction (Forward or Reverse), Average Flow and Average Sound Velocities values. b. Click the Path Information tab and view the Chord performance, Gain, SNR (Signal to Noise Ratio) Signal strength (mV), and Noise (mV). c. Click the Flow Totals tab to view the volume totals (forward and reverse uncorrected volume).

82 Using AMS Device Manager to configure the meter

Page 97 of the Emerson 3812 user manual — 4.3 Using AMS Device Manager to configure the meter
Page 97 · click to enlarge
Page 97

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 4: Configuration 3-9000-765 Rev H May 2015

d. Click the All Variables tab to view a graphical display of the Primary, Secondary, Third and Fourth Variables.

Figure 4-6 AMS Device Manager - Service Tools All Variables status indicators

Emerson 3812 — 4.3 Using AMS Device Manager to configure the meter diagram (page 97)
Fig. 47 · 4.3 Using AMS Device Manager to configure the meter
8. Click OK to return to the Overview page.
9. Enable the Write Protect switch on the CPU Module to protect the meter’s

configuration.

10. From the Overview window, click Display Meter K-Factors . K-Factors are a read-only

values calculated from the Full scale volumetric flow rate used with frequency outputs and the Maximum frequency for frequency output.

Figure 4-7 Display Meter K-Factors

Emerson 3812 — 4.3 Using AMS Device Manager to configure the meter diagram (page 97)
Fig. 48 · 4.3 Using AMS Device Manager to configure the meter

Click Next to return to the Device Manager Overview page.

Using AMS Device Manager to configure the meter 83

Page 98 of the Emerson 3812 user manual — 4.3 Using AMS Device Manager to configure the meter
Page 98 · click to enlarge
Page 98
Section 4Configuration Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

AMS Device Manager - Manual Setup

Use the Manual Setup wizard to configure the meter’s parameters. See Figure 4-3 and Figure 4-4 and from the AMS Device Manager Configure menu click Manual Setup.

Figure 4-8 AMS Device Manager - Configure Manual Setup

Emerson 3812 — 4.3 Using AMS Device Manager to configure the meter diagram (page 98)
Fig. 49 · 4.3 Using AMS Device Manager to configure the meter
1. If installed, remove security wires from the endcap and the Bracket/Cover hex head bolts that secures the Base Enclosure.
1. Disable the Write Protect switch in the CPU Module to write any of the following configuration parameters to your meter.
2. Click the Device Variables Mapping tab. The Primary and Secondary variables are read

only and are configured for Uncorrected Flow Rate. The Third and Fourth variable configuration choices include Pressure and Temperature.

3. Click the Units tab (see AMS Device Manager - Guided Setup , Step 1 ).
4. Click the Analog Output 1 (HART) tab (see AMS Device Manager - Guided Setup, Step

2a.).

5. Click the Analog Output 2 tab. Follow the configuration instructions in the AMS Device

Manager - Guided Setup, Step 2b. The read only Secondary variable Content, Uncorrected Flow Rate, displays. Use the drop-down arrow and select the (flow) Direction - Forward or Reverse. Enter a Lower and Upper Range limit. Set the Alarm Action parameters. Click Apply, after you enter the data to write the parameters to the meter.

6. Click the Frequency/Digital Outputs tab. Follow the configuration instructions in the

AMS Device Manager - Guided Setup, Step 3 a.). (Refresh Note: If changes are made to any Source variable on this page, apply the changes and navigate to the Guided Setup

84 Using AMS Device Manager to configure the meter

Procedure

Page 99 of the Emerson 3812 user manual — 4.3 Using AMS Device Manager to configure the meter
Page 99 · click to enlarge
Page 99

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 4: Configuration 3-9000-765 Rev H May 2015

page. Navigate back to the Manual Setup for the changes to be reflected in other Manual Setup pages). Click Apply, after you enter the data to write the parameters to the meter.

7. Click the Frequency and Digital Output 1 tab. Follow the configuration instructions in the

AMS Device Manager - Guided Setup, Step 3b. Click Apply, after you enter the data to write the parameters to the meter.

8. Click the Frequency and Digital Output 2 tab. Follow the instructions in the AMS Device

Manager - Guided Setup, Step 3c to configure the Frequency and Digital Output 2 parameters. Click Apply, after you enter the data to write the parameters to the meter.

9. Click the Temperature tab. Configure the input parameters including: Source (Live

Analog or Fixed), Min and Max input limits corresponding to 4 mA and 20 mA respectively and the Low and High alarm limits. Click Apply, after you enter the data to write the parameters to the meter.

10. Click the Pressure tab. Configure the input parameters including: Source (Live Analog or

Fixed), Min and Max input limits corresponding to 4 mA and 20 mA respectively and the Low and High alarm limits. Select either Gage or Absolute for the type of pressure reading desired. If a live pressure transmitter is connected, select the type of reading the transmitter outputs. If Absolute is selected, you must also enter the Atmospheric pressure. Click Apply, after you enter the data to write the parameters to the meter.

11. Click the Digital Input tab. The default Digital Input 1 polarity is set to Normal for general

purpose or set to Inverted when used for calibration. Click Apply, after you choose the calibration data to write the parameters to the meter.

a. Calibration Polarity configuration parameter selections are: – Digital Input 1 Calibrate Active High – Digital Input 1 Calibrate Active Low

Using AMS Device Manager to configure the meter 85

Page 100 of the Emerson 3812 user manual — 4.3 Using AMS Device Manager to configure the meter
Page 100 · click to enlarge
Page 100
Section 4Configuration Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

b. Calibration Gating configuration parameter selections are: – Edge gated, active high

Figure 4-9 Gating configuration parameter Edge gated, active high

Calibration Start Calibration Stop

– Edge gated, active low

Figure 4-10 Gating configuration parameter Edge gated, active low

Calibration Start Calibration Stop

– State gated, active high

Figure 4-11 Gating configuration parameter State gated, active high

Calibration Start Calibration Stop

86 Using AMS Device Manager to configure the meter

Show 6 more labels / short text on this page
High · Low · High · Low · High · Low
Page 101 of the Emerson 3812 user manual — 4.3 Using AMS Device Manager to configure the meter
Page 101 · click to enlarge
Page 101

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 4: Configuration 3-9000-765 Rev H May 2015

– State gated, active low

Figure 4-12 Gating configuration parameter State gated, active low

Calibration Start Calibration Stop

12. Click the Alert Setup tab (from the main Configuration page).

Figure 4-13 Configure Flow Analysis Alert

Emerson 3812 — 4.3 Using AMS Device Manager to configure the meter diagram (page 101)
Fig. 50 · 4.3 Using AMS Device Manager to configure the meter

Using AMS Device Manager to configure the meter 87

High · Low

Page 102 of the Emerson 3812 user manual — 4.3 Using AMS Device Manager to configure the meter
Page 102 · click to enlarge
Page 102
Section 4Configuration Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H
13. Click the Flow Analysis tab to select Configure Reverse Flow Detection, if desired. The

default setting is Disabled. Click the Disabled button to send the feature command to the meter. Check for a response error. If no error response is received, click the Enable button.

a. Enter the minimum reverse flow velocity above which to accumulate flow in the reverse direction for this alert. Enter a positive value for the Reverse Flow Zero Cutoff. Click the Next button to write the values to the meter. Check for an error response. If no error response is received, click the Next button. The Detect Reverse Flow enabled page displays. Click the Next button to display Detect Reverse Flow disabled. b. If an error message is returned, click the Next button to display the Method Complete page. c. Click the Set Flow Range Limits button and enter a positive value for the Flow Analysis Lower Velocity Range and the Upper Velocity Range Limits. When the velocity is outside of the limit parameters, an alert is triggered. Click the Next button to display the Method Complete page.

88 Using AMS Device Manager to configure the meter

Page 103 of the Emerson 3812 user manual — 4.3 Using AMS Device Manager to configure the meter
Page 103 · click to enlarge
Page 103

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 4: Configuration 3-9000-765 Rev H May 2015

14. Click the Service Tools tab to access the device alerts, variables, trends and maintenance

statuses or to edit the configuration parameters.

a. Click the Service Tools|Alerts tab. If an alert condition exists, the alert type and description displays. Recommended actions are listed to assist you in a resolution. After you resolve the alert condition, click the Acknowledge button to clear the alert. Click Apply to write the changes to the meter. If no alert condition is active, click OK to close the device window.

Figure 4-14 AMS Device Manager - Service Tools Alerts

Emerson 3812 — 4.3 Using AMS Device Manager to configure the meter diagram (page 103)
Fig. 51 · 4.3 Using AMS Device Manager to configure the meter

b. If you change the device configuration, a confirmation dialog displays and prompts you to write the changes to the meter. Click Yes to write the changes to the meter or click No to cancel pending changes.

Figure 4-15 Configuration changes dialog

Emerson 3812 — 4.3 Using AMS Device Manager to configure the meter diagram (page 103)
Fig. 52 · 4.3 Using AMS Device Manager to configure the meter

Using AMS Device Manager to configure the meter 89

Page 104 of the Emerson 3812 user manual — 4.3 Using AMS Device Manager to configure the meter
Page 104 · click to enlarge
Page 104
Section 4Configuration Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

c. Click the Service Tools|Variables tab. The Variables page displays tabs for the device’s Flow Data, Path Information, Flow Totals, and All Variables).

Figure 4-16 AMS Device Manager - Service Tools

Emerson 3812 — 4.3 Using AMS Device Manager to configure the meter diagram (page 104)
Fig. 53 · 4.3 Using AMS Device Manager to configure the meter

d. The Service Tools|Flow Data page includes charts for flow and sound velocities. The flow values (flow direction, average flow velocity and average sound velocity) parameters are displayed for the connected device. e. Click Service Tools|Variables|Path Information tab to view the device’s chord performance (%), Gain (dB), SNR (dB), Signal (mV) and Noise (mV). f. Click Service Tools|Variables|Flow Totals to view the volume totals (Forward and reverse Uncorrected Volume) parameters for the connected device.

90 Using AMS Device Manager to configure the meter

Page 105 of the Emerson 3812 user manual — IMPORTANT
Page 105 · click to enlarge
Page 105

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 4: Configuration 3-9000-765 Rev H May 2015

g. Click Service Tools|Variables|All Variables tab to view Primary, Secondary, Third and Fourth Variable parameter status.

Figure 4-17 AMS Device Manager - Service Tools All Variables

Gauges display each variable’s status as good or bad. If a status is bad refer to the Service Tools Alerts page for recommended actions to resolve the alert condition. Also refer to the Field Device Specification manual (P/N 3-9000-762) for Commands 48 and 140 details.

IMPORTANT

Alerts are triggered for Command 48 Additional device status and Command 140 detailed status information. Alerts are grouped as Failed - Fix Now, Maintenance - Fix Soon and Advisory according to the severity level; 1-6. Severity 1 is the highest and 6 is the lowest level.

Using AMS Device Manager to configure the meter 91

Page 106 of the Emerson 3812 user manual — 4.3 Using AMS Device Manager to configure the meter
Page 106 · click to enlarge
Page 106
Section 4Configuration Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

h. Click the Service Tools|Trends tab to display the device variables (uncorrected volume flow rate, pressure and temperature) trends.

Figure 4-18 AMS Device Manager - Service Tools Trends

Emerson 3812 — 4.3 Using AMS Device Manager to configure the meter diagram (page 106)
Fig. 54 · 4.3 Using AMS Device Manager to configure the meter

Primary and Secondary variables display real-time uncorrected volume flow rate trends. The third and fourth variables charts displays trends for temperature and pressure.

15. Click the Service Tools|Routine Maintenance tab. Click Analog Output 1 Trim to perform a

digital to analog trim adjustment of the first milliampere output. The 4mA and 20mA output current values should equal the plant’s standard values. Click Yes to confirm the configuration changes. Repeat this step to trim Analog Output 2 current. Click Apply to write the output trim values to the meter. Click OK to navigate back to the Service Tools page.

16. Click the Service Tools|Zero Calibration tab. See AMS Device Manager - Guided Setup

Step 6 to configure the zero flow parameters.

17. After you have changed and written the configuration changes to the meter do the

following:

a. Enable the Write Protect switch on the CPU Module to protect the meter’s configuration. b. Replace the end cap and if required, apply security seals through the endcap holes and through the hex head bolts that secure the Bracket/Cover to the Base enclosure. Note: The next time you connect to the device using Daniel MeterLink, the Monitor page displays a Meter status alarm that the configuration has changed and remains latched until acknowledged. Click Ack (acknowledge) to clear the alarm.

92 Using AMS Device Manager to configure the meter

Page 107 of the Emerson 3812 user manual — 4.4 Using a Field Communicator to configure the
Page 107 · click to enlarge
Page 107

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 4: Configuration 3-9000-765 Rev H May 2015

4.4 Using a Field Communicator to configure the

meter

Follow all guidelines and precautions described in the Field Communicator User Manual and in the 3812 Liquid Ultrasonic Flow Meter documentation when working in a hazardous area.

  • Emerson Field Communicator software, license, installation guide and user manual available on the Emerson Asset Optimization Field Communicator website:

http://www2.emersonprocess.com/en-US/brands/Field-Communicator/Pages/Documenta tion.aspx

  • Daniel HART Device Description (HART DD) installed for the meter
  • Network configured for a Field Communicator
  • Daniel Field Device Specification Manual (P/N 3-9000-761) available on the Daniel website

http://www2.emersonprocess.com/en-US/brands/daniel/Flow/ultrasonics/Pages/Ultra sonic.aspx

  • System wiring diagram drawing number DMC - 004936 ( see Appendix A )
  • Power supply

Using a Field Communicator to configure the meter 93

Installation Requirements

Page 108 of the Emerson 3812 user manual — 4.4 Using a Field Communicator to configure the meter
Page 108 · click to enlarge
Page 108
Section 4Configuration Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H
1. Remove electrical power to the meter. If installed, remove the endcap security latches and seals and then, remove the endcap.
2. Refer to the Field Communicator Users Manual wiring diagrams and commissioning

instructions provided with your handheld device. Register the product to activate the end user license.

3. Fully charge the Field Communicator battery prior to use. Important: Do not change

the battery in a hazardous area environment. The power supply is not intrinsically safe.

4. On the meter, run the wires through the field wiring conduit and into the transmitter

electronics enclosure.

Figure 4-19 3812 transmitter field wiring conduit entries

Emerson 3812 — 4.4 Using a Field Communicator to configure the meter diagram (page 108)
Fig. 55 · 4.4 Using a Field Communicator to configure the mete…

A.

A. Field wiring conduit entries (4)

94 Using a Field Communicator to configure the meter

Procedure

Page 109 of the Emerson 3812 user manual — 4.4 Using a Field Communicator to configure the meter
Page 109 · click to enlarge
Page 109

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 4: Configuration 3-9000-765 Rev H May 2015

5. Wire Analog Input 1 (AI1) and Analog Output 1 (AO1) as shown in Figure 4-20 and

Appendix A, drawing DMC-004936.

Figure 4-20 Field Communicator wiring diagram for the 3810 Series electronics

Emerson 3812 — 4.4 Using a Field Communicator to configure the meter diagram (page 109)
Fig. 56 · 4.4 Using a Field Communicator to configure the mete…
Emerson 3812 — 4.4 Using a Field Communicator to configure the meter diagram (page 109)
Fig. 57 · 4.4 Using a Field Communicator to configure the mete…
6. Use the leads provided with the Field Communicator to connect to your device.
7. Press and hold the Power button on the Field Communicator until the green light

blinks.

8. Use the touch screen on the Field Communicator, the keypad or use the stylus to

navigate through the device menus.

9. Refer to the Menu tree in Section D.1.1 of the Daniel HART Field Device Specification

manual (P/N 3-9000-761) for the device fast key sequences. Included in the menu tree are:

  • Diagram Page 1 - 3810 Series Root Menu; Overview, Configure|Manual Setup
  • Diagram Page 2 - Configure|Manual Setup (continued) and Alerts Setup
  • Diagram Page 3 - Service Tools|Alerts and Variables
  • Diagram Page 4 - Service Tools|Variables (continued), Service Tools|Trends , and Service Tools|Maintenance

If you encounter problems, refer to the contact information on the back cover of this manual or the contacts included in the Field Communicator Users Manual.

Using a Field Communicator to configure the meter 95

Page 110 of the Emerson 3812 user manual — 4.5 Security seals for the meter
Page 110 · click to enlarge
Page 110
Section 4Configuration Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

4.5 Security seals for the meter

For the integrity of the meter metrology and to prevent tampering with the transmitter electronics and transducer assemblies, attach security latches to the Shroud covering the transducer assemblies and install security wires on the Transmitter Electronics Enclosure end caps, the Base Enclosure hex head bolts, the Shroud clamps and the shroud latches (see Section 3.7).

Pour the conduit ports sealing compound according to the customer’s requirements (e.g., after approximately one to two weeks of run time). Also, see Section 3.4.1 and see Section 3.4.2.

96 Security seals for the meter

Page 111 of the Emerson 3812 user manual — Appendix A Appendix A: Engineering drawings
Page 111 · click to enlarge
Page 111

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Engineering drawings 3-9000-765 Rev H May 2015

Appendix A Appendix A: Engineering drawings

A.1 Daniel 3812 Liquid Ultrasonic Flow Meter drawings

List of Engineering Drawings

This appendix contains the following engineering drawing(s) for the ultrasonic meter:

DMC-004936 Daniel 3810 Series Ultrasonic Flow Meter System Wiring Diagram

Daniel 3812 Liquid Ultrasonic Flow Meter drawings 97

Page 112 of the Emerson 3812 user manual — Appendix A: Engineering drawings
Page 112 · click to enlarge
Page 112

Engineering drawings Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

98 Daniel 3812 Liquid Ultrasonic Flow Meter drawings

Page 113 of the Emerson 3812 user manual — Appendix A: Engineering drawings
Page 113 · click to enlarge
Page 113
Emerson 3812 — Appendix A: Engineering drawings diagram (page 113)
Fig. 58 · Appendix A: Engineering drawings
Page 114 of the Emerson 3812 user manual — Appendix A: Engineering drawings
Page 114 · click to enlarge
Page 114
Emerson 3812 — Appendix A: Engineering drawings diagram (page 114)
Fig. 59 · Appendix A: Engineering drawings
Page 115 of the Emerson 3812 user manual — Appendix A: Engineering drawings
Page 115 · click to enlarge
Page 115
Emerson 3812 — Appendix A: Engineering drawings diagram (page 115)
Fig. 60 · Appendix A: Engineering drawings
Page 116 of the Emerson 3812 user manual — Appendix A: Engineering drawings
Page 116 · click to enlarge
Page 116
Page 116 of the Emerson 3812 user manual — Appendix A: Engineering drawings
Page 117 of the Emerson 3812 user manual — A Appendix B: Open source licenses
Page 117 · click to enlarge
Page 117

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Open source licenses 3-9000-765 Rev H May 2015

A Appendix B: Open source licenses

Source code for executable files or libraries included in this product is provided per the indicated license in the table below. Hyperlinks to the controlling organization's websites are included in Section B.1 through Section B.4.

Table B-1 Open source licences

Package File specification License Summary

base_libs-1.2-1 base_libs LGPL Base Libraries (from toolchain)

busybox-1.1.3-1 busybox GPL A small executable that replaces many UNIX utilities

dev-1.1-1 dev GPL Device files for a small embedded system

devmem2-1.0-1 devmem2 GPL Simple program to read/write from/to any location

ethtool-3-1 ethtool GPL Ethernet settings tool for PCI Ethernet cards

fake-provides-1.0-5 fake-provides GPL Fake provides to satisfy package dependencies

gdb-6.6cs-1 gdb GPL Gdb - GNU Source level debugger for C, C++

kernel-2.6.37-6 kernel-2.6.37- mpc8313erd

GPL Linux kernel (core of the Linux operating system)

libpcap-0.8.3-1 libpcap BSD A system-independent interface for user-level pa

libtermcap-2.0.8-31_1 libtermcap LGPL A basic system library for accessing the termcap

lwIP BSD A lightweight TCP/IP stack

merge-0.1-1 merge GPL Merge files for an embedded root filesystem

modeps-1.0-1 modeps GPL Generate module dependency file

mtd-utils-20060302-1 mtd-utils GPL Memory Technology Device tools

net-tools-1.60-1 net-tools GPL Basic networking tools

ppp-2.4.4-1 ppp BSD Like a Point-to-Point Protocol daemon

skell-1.16-2 skell GPL Skelleton files for an embedded root filesystem

sqlite-3.6.22-1 sqlite Public domain SQLite is a C library that imple ments an embeddable SQL database

Page 118 of the Emerson 3812 user manual — http://www2.emersonprocess.com/en-US/brands/daniel/Pages/GPL3810.aspx B.1 GNU General Public License
Page 118 · click to enlarge
Page 118

Open source licenses Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

Package File specification License Summary

strace-4.5.14-1 strace BSD trace system calls associated with a running pro

sysconfig-1.2-1 sysconfig GPL System configuration package

sysfsutils-2.1.0-1 sysfsutils GPL/LGPL sysfs utilities

tcpdump-3.8.3-1 tcpdump BSD A network traffic monitoring tool

termcap-1.2-1 termcap BSD minimal /etc/termcap needed by minicom etc

u-boot-1.3.0-1 u-boot-1.3.0- mpc8313erdb

GPL Universal Bootloader firmware

ubi-utils-1.4.2-1 ubi-utils GPL Tools for maintaining Unsorted Block Image Device

vsftpd-2.2.2-1 vsftpd GPL vsftpd - Very Secure Ftp Daemon

zlib-1.2.3-2 zlib zlib Distribution zlib compression utilities and libraries

Follow the link below to the Daniel® Liquid Ultrasonic Products GPL webpage for additional open source information and zipped source code files.

http://www2.emersonprocess.com/en-US/brands/daniel/Pages/GPL3810.aspx B.1 GNU General Public License

For more details about GNU GPL (General Public License), follow the link below:

http://www.gnu.org/

Daniel® Measurement and Control, Inc., uses GPL version 2.

http://www.gnu.org/licenses/old-licenses/gpl-2.0.html

The GNU GPL is currently version 3

http://www.gnu.org/licenses/quick-guide-gplv3.html

For older versions of the GNU General Public License, follow the link below:

http://www.gnu.org/licenses/old-licenses/old-licenses.html#GPL

See GPL license on the following pages.

100 GNU General Public License

Page 119 of the Emerson 3812 user manual — Version 2, June 1991
Page 119 · click to enlarge
Page 119

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Open source licenses 3-9000-765 Rev H May 2015

The GNU General Public License (GPL)

Version 2, June 1991

Copyright (C) 1989, 1991 Free Software Foundation, Inc.

59 Temple Place, Suite 330, Boston, MA 02111-1307 USA

Everyone is permitted to copy and distribute verbatim copies

of this license document, but changing it is not allowed.

The licenses for most software are designed to take away your freedom to share and change it. By

contrast, the GNU General Public License is intended to guarantee your freedom to share and change

free software--to make sure the software is free for all its users. This General Public License applies to

most of the Free Software Foundation's software and to any other program whose authors commit to

using it. (Some other Free Software Foundation software is covered by the GNU Library General Public

License instead.) You can apply it to your programs, too.

When we speak of free software, we are referring to freedom, not price. Our General Public Licenses

are designed to make sure that you have the freedom to distribute copies of free software (and

charge for this service if you wish), that you receive source code or can get it if you want it, that you

can change the software or use pieces of it in new free programs; and that you know you can do

these things.

To protect your rights, we need to make restrictions that forbid anyone to deny you these rights or to

ask you to surrender the rights. These restrictions translate to certain responsibilities for you if you

distribute copies of the software, or if you modify it.

GNU General Public License 101

Preamble

Page 120 of the Emerson 3812 user manual — B.1 GNU General Public License
Page 120 · click to enlarge
Page 120

Open source licenses Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

For example, if you distribute copies of such a program, whether gratis or for a fee, you must give the

recipients all the rights that you have. You must make sure that they, too, receive or can get the

source code. And you must show them these terms so they know their rights.

We protect your rights with two steps: (1) copyright the software, and (2) offer you this license which

gives you legal permission to copy, distribute and/or modify the software.

Also, for each author's protection and ours, we want to make certain that everyone understands that

there is no warranty for this free software. If the software is modified by someone else and passed

on, we want its recipients to know that what they have is not the original, so that any problems

introduced by others will not reflect on the original authors' reputations.

Finally, any free program is threatened constantly by software patents. We wish to avoid the danger

that redistributors of a free program will individually obtain patent licenses, in effect making the

program proprietary. To prevent this, we have made it clear that any patent must be licensed for

everyone's free use or not licensed at all.

The precise terms and conditions for copying, distribution and modification follow.

102 GNU General Public License

Page 121 of the Emerson 3812 user manual — TERMS AND CONDITIONS FOR COPYING, DISTRIBUTION AND MODIFICATION
Page 121 · click to enlarge
Page 121

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Open source licenses 3-9000-765 Rev H May 2015

TERMS AND CONDITIONS FOR COPYING, DISTRIBUTION AND MODIFICATION

0. This License applies to any program or other work which contains a notice placed by the copyright

holder saying it may be distributed under the terms of this General Public License. The "Program",

below, refers to any such program or work, and a "work based on the Program" means either the

Program or any derivative work under copyright law: that is to say, a work containing the Program or

a portion of it, either verbatim or with modifications and/or translated into another language.

(Hereinafter, translation is included without limitation in the term "modification".) Each licensee is

addressed as "you".

Activities other than copying, distribution and modification are not covered by this License; they are

outside its scope. The act of running the Program is not restricted, and the output from the Program

is covered only if its contents constitute a work based on the Program (independent of having been

made by running the Program). Whether that is true depends on what the Program does.

1. You may copy and distribute verbatim copies of the Program's source code as you receive it, in any

medium, provided that you conspicuously and appropriately publish on each copy an appropriate

copyright notice and disclaimer of warranty; keep intact all the notices that refer to this License and

to the absence of any warranty; and give any other recipients of the Program a copy of this License

along with the Program.

You may charge a fee for the physical act of transferring a copy, and you may at your option offer

warranty protection in exchange for a fee.

GNU General Public License 103

Page 122 of the Emerson 3812 user manual — B.1 GNU General Public License
Page 122 · click to enlarge
Page 122

Open source licenses Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

2. You may modify your copy or copies of the Program or any portion of it, thus forming a work based

on the Program, and copy and distribute such modifications or work under the terms of Section 1

above, provided that you also meet all of these conditions:

a) You must cause the modified files to carry prominent notices stating that you changed the

files and the date of any change.

b) You must cause any work that you distribute or publish, that in whole or in part contains or

is derived from the Program or any part thereof, to be licensed as a whole at no charge to

all third parties under the terms of this License.

c) If the modified program normally reads commands interactively when run, you must cause

it, when started running for such interactive use in the most ordinary way, to print or

display an announcement including an appropriate copyright notice and a notice that there

is no warranty (or else, saying that you provide a warranty) and that users may

redistribute the program under these conditions, and telling the user how to view a copy

of this License. (Exceptionif the Program itself is interactive but does not normally print

such an announcement, your work based on the Program is not required to print an

announcement.)

These requirements apply to the modified work as a whole. If identifiable sections of that work are not

derived from the Program, and can be reasonably considered independent and separate works in

themselves, then this License, and its terms, do not apply to those sections when you distribute them

as separate works. But when you distribute the same sections as part of a whole which is a work

based on the Program, the distribution of the whole must be on the terms of this License, whose

permissions for other licensees extend to the entire whole, and thus to each and every part regardless

of who wrote it.

Thus, it is not the intent of this section to claim rights or contest your rights to work written entirely

by you; rather, the intent is to exercise the right to control the distribution of derivative or collective

works based on the Program.

104 GNU General Public License

Page 123 of the Emerson 3812 user manual — B.1 GNU General Public License
Page 123 · click to enlarge
Page 123

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Open source licenses 3-9000-765 Rev H May 2015

In addition, mere aggregation of another work not based on the Program with the Program (or with a

work based on the Program) on a volume of a storage or distribution medium does not bring the other

work under the scope of this License.

3. You may copy and distribute the Program (or a work based on it, under Section 2) in object code or

executable form under the terms of Sections 1 and 2 above provided that you also do one of the

following:

a) Accompany it with the complete corresponding machine-readable source code, which must

be distributed under the terms of Sections 1 and 2 above on a medium customarily used

for software interchange; or,

b) Accompany it with a written offer, valid for at least three years, to give any third party, for

a charge no more than your cost of physically performing source distribution, a complete

machine-readable copy of the corresponding source code, to be distributed under the

terms of Sections 1 and 2 above on a medium customarily used for software interchange;

c) Accompany it with the information you received as to the offer to distribute corresponding

source code. (This alternative is allowed only for noncommercial distribution and only if

you received the program in object code or executable form with such an offer, in accord

with Subsection b above.)

The source code for a work means the preferred form of the work for making modifications to it. For

an executable work, complete source code means all the source code for all modules it contains, plus

any associated interface definition files, plus the scripts used to control compilation and installation of

the executable. However, as a special exception, the source code distributed need not include

anything that is normally distributed (in either source or binary form) with the major components

(compiler, kernel, and so on) of the operating system on which the executable runs, unless that

component itself accompanies the executable.

GNU General Public License 105

or,

Page 124 of the Emerson 3812 user manual — B.1 GNU General Public License
Page 124 · click to enlarge
Page 124

Open source licenses Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

If distribution of executable or object code is made by offering access to copy from a designated

place, then offering equivalent access to copy the source code from the same place counts as

distribution of the source code, even though third parties are not compelled to copy the source along

with the object code.

4. You may not copy, modify, sublicense, or distribute the Program except as expressly provided

under this License. Any attempt otherwise to copy, modify, sublicense or distribute the Program is

void, and will automatically terminate your rights under this License. However, parties who have

received copies, or rights, from you under this License will not have their licenses terminated so long

as such parties remain in full compliance.

5. You are not required to accept this License, since you have not signed it. However, nothing else

grants you permission to modify or distribute the Program or its derivative works. These actions are

prohibited by law if you do not accept this License. Therefore, by modifying or distributing the

Program (or any work based on the Program), you indicate your acceptance of this License to do so,

and all its terms and conditions for copying, distributing or modifying the Program or works based on

it.

6. Each time you redistribute the Program (or any work based on the Program), the recipient

automatically receives a license from the original licensor to copy, distribute or modify the Program

subject to these terms and conditions. You may not impose any further restrictions on the recipients'

exercise of the rights granted herein. You are not responsible for enforcing compliance by third

parties to this License.

7. If, as a consequence of a court judgment or allegation of patent infringement or for any other

reason (not limited to patent issues), conditions are imposed on you (whether by court order,

agreement or otherwise) that contradict the conditions of this License, they do not excuse you from

the conditions of this License. If you cannot distribute so as to satisfy simultaneously your obligations

under this License and any other pertinent obligations, then as a consequence you may not distribute

the Program at all. For example, if a patent license would not permit royalty-free redistribution of the

106 GNU General Public License

Page 125 of the Emerson 3812 user manual — B.1 GNU General Public License
Page 125 · click to enlarge
Page 125

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Open source licenses 3-9000-765 Rev H May 2015

Program by all those who receive copies directly or indirectly through you, then the only way you

could satisfy both it and this License would be to refrain entirely from distribution of the Program.

If any portion of this section is held invalid or unenforceable under any particular circumstance, the

balance of the section is intended to apply and the section as a whole is intended to apply in other

circumstances.

It is not the purpose of this section to induce you to infringe any patents or other property right

claims or to contest validity of any such claims; this section has the sole purpose of protecting the

integrity of the free software distribution system, which is implemented by public license practices.

Many people have made generous contributions to the wide range of software distributed through

that system in reliance on consistent application of that system; it is up to the author/donor to decide

if he or she is willing to distribute software through any other system and a licensee cannot impose

that choice.

This section is intended to make thoroughly clear what is believed to be a consequence of the rest of

this License.

8. If the distribution and/or use of the Program is restricted in certain countries either by patents or

by copyrighted interfaces, the original copyright holder who places the Program under this License

may add an explicit geographical distribution limitation excluding those countries, so that distribution

is permitted only in or among countries not thus excluded. In such case, this License incorporates the

limitation as if written in the body of this License.

9. The Free Software Foundation may publish revised and/or new versions of the General Public

License from time to time. Such new versions will be similar in spirit to the present version, but may

differ in detail to address new problems or concerns.

Each version is given a distinguishing version number. If the Program specifies a version number of

this License which applies to it and "any later version", you have the option of following the terms and

GNU General Public License 107

Page 126 of the Emerson 3812 user manual — END OF TERMS AND CONDITIONS
Page 126 · click to enlarge
Page 126

Open source licenses Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

conditions either of that version or of any later version published by the Free Software Foundation. If

the Program does not specify a version number of this License, you may choose any version ever

published by the Free Software Foundation.

10. If you wish to incorporate parts of the Program into other free programs whose distribution

conditions are different, write to the author to ask for permission. For software which is copyrighted

by the Free Software Foundation, write to the Free Software Foundation; we sometimes make

exceptions for this. Our decision will be guided by the two goals of preserving the free status of all

derivatives of our free software and of promoting the sharing and reuse of software generally.

NO WARRANTY

11. BECAUSE THE PROGRAM IS LICENSED FREE OF CHARGE, THERE IS NO WARRANTY FOR THE

PROGRAM, TO THE EXTENT PERMITTED BY APPLICABLE LAW. EXCEPT WHEN OTHERWISE STATED IN

WRITING THE COPYRIGHT HOLDERS AND/OR OTHER PARTIES PROVIDE THE PROGRAM "AS IS"

WITHOUT WARRANTY OF ANY KIND, EITHER EXPRESSED OR IMPLIED, INCLUDING, BUT NOT

LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR

PURPOSE. THE ENTIRE RISK AS TO THE QUALITY AND PERFORMANCE OF THE PROGRAM IS WITH

YOU. SHOULD THE PROGRAM PROVE DEFECTIVE, YOU ASSUME THE COST OF ALL NECESSARY

SERVICING, REPAIR OR CORRECTION.

12. IN NO EVENT UNLESS REQUIRED BY APPLICABLE LAW OR AGREED TO IN WRITING WILL ANY

COPYRIGHT HOLDER, OR ANY OTHER PARTY WHO MAY MODIFY AND/OR REDISTRIBUTE THE

PROGRAM AS PERMITTED ABOVE, BE LIABLE TO YOU FOR DAMAGES, INCLUDING ANY GENERAL,

SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES ARISING OUT OF THE USE OR INABILITY TO

USE THE PROGRAM (INCLUDING BUT NOT LIMITED TO LOSS OF DATA OR DATA BEING RENDERED

INACCURATE OR LOSSES SUSTAINED BY YOU OR THIRD PARTIES OR A FAILURE OF THE PROGRAM

TO OPERATE WITH ANY OTHER PROGRAMS), EVEN IF SUCH HOLDER OR OTHER PARTY HAS BEEN

ADVISED OF THE POSSIBILITY OF SUCH DAMAGES.

END OF TERMS AND CONDITIONS

108 GNU General Public License

Page 127 of the Emerson 3812 user manual — 02111-1307 USA
Page 127 · click to enlarge
Page 127

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Open source licenses 3-9000-765 Rev H May 2015

How to Apply These Terms to Your New Programs

If you develop a new program, and you want it to be of the greatest possible use to the public, the

best way to achieve this is to make it free software which everyone can redistribute and change under

these terms.

To do so, attach the following notices to the program. It is safest to attach them to the start of each

source file to most effectively convey the exclusion of warranty; and each file should have at least the

"copyright" line and a pointer to where the full notice is found.

One line to give the program's name and a brief idea of what it does.

Copyright (C)

This program is free software; you can redistribute it and/or modify it under the terms of the

GNU General Public License as published by the Free Software Foundation; either version 2 of

the License, or (at your option) any later version.

This program is distributed in the hope that it will be useful, but WITHOUT ANY WARRANTY;

without even the implied warranty of MERCHANTABILITY or FITNESS FOR A PARTICULAR

PURPOSE. See the GNU General Public License for more details.

You should have received a copy of the GNU General Public License along with this program;

if not, write to the Free Software Foundation, Inc., 59 Temple Place, Suite 330, Boston, MA

02111-1307 USA

Also add information on how to contact you by electronic and paper mail.

If the program is interactive, make it output a short notice like this when it starts in an interactive

mode:

Gnomovision version 69, Copyright (C) year name of author Gnomovision comes with

ABSOLUTELY NO WARRANTY; for details type `show w'. This is free software, and you are

welcome to redistribute it under certain conditions; type `show c' for details.

GNU General Public License 109

Page 128 of the Emerson 3812 user manual — B.1 GNU General Public License
Page 128 · click to enlarge
Page 128

Open source licenses Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

The hypothetical commands `show w' and `show c' should show the appropriate parts of the General

Public License. Of course, the commands you use may be called something other than `show w' and

`show c'; they could even be mouse-clicks or menu items--whatever suits your program.

You should also get your employer (if you work as a programmer) or your school, if any, to sign a

"copyright disclaimer" for the program, if necessary. Here is a sample; alter the names:

Yoyodyne, Inc., hereby disclaims all copyright interest in the program `Gnomovision' (which

makes passes at compilers) written by James Hacker.

signature of Ty Coon, 1 April 1989

Ty Coon, President of Vice

This General Public License does not permit incorporating your program into proprietary

programs. If your program is a subroutine library, you may consider it more useful to permit

linking proprietary applications with the library. If this is what you want to do, use the GNU

Library General Public License instead of this License.

110 GNU General Public License

Page 129 of the Emerson 3812 user manual — GNU LESSER GENERAL PUBLIC LICENSE
Page 129 · click to enlarge
Page 129

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Open source licenses 3-9000-765 Rev H May 2015

B.2 GNU Lesser General Public License

GNU LESSER GENERAL PUBLIC LICENSE

Version 3, 29 June 2007

Copyright © 2007 Free Software Foundation, Inc.

Everyone is permitted to copy and distribute verbatim copies of this license document, but

changing it is not allowed.

This version of the GNU Lesser General Public License incorporates the terms and conditions of

version 3 of the GNU General Public License, supplemented by the additional permissions listed

below.

0. Additional Definitions.

As used herein, "this License" refers to version 3 of the GNU Lesser General Public License, and

the "GNU GPL" refers to version 3 of the GNU General Public License.

"The Library" refers to a covered work governed by this License, other than an Application or a

Combined Work as defined below.

An "Application" is any work that makes use of an interface provided by the Library, but which is

not otherwise based on the Library. Defining a subclass of a class defined by the Library is

deemed a mode of using an interface provided by the Library.

A "Combined Work" is a work produced by combining or linking an Application with the Library.

The particular version of the Library with which the Combined Work was made is also called the

"Linked Version".

The "Minimal Corresponding Source" for a Combined Work means the Corresponding Source for

the Combined Work, excluding any source code for portions of the Combined Work that,

considered in isolation, are based on the Application, and not on the Linked Version.

GNU Lesser General Public License 111

Page 130 of the Emerson 3812 user manual — B.2 GNU Lesser General Public License
Page 130 · click to enlarge
Page 130

Open source licenses Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

The "Corresponding Application Code" for a Combined Work means the object code and/or

source code for the Application, including any data and utility programs needed for reproducing

the Combined Work from the Application, but excluding the System Libraries of the Combined

Work.

1. Exception to Section 3 of the GNU GPL.

You may convey a covered work under sections 3 and 4 of this License without being bound by

section 3 of the GNU GPL.

2. Conveying Modified Versions.

If you modify a copy of the Library, and, in your modifications, a facility refers to a function or

data to be supplied by an Application that uses the facility (other than as an argument passed

when the facility is invoked), then you may convey a copy of the modified version:

"a) under this License, provided that you make a good faith effort to ensure that, in the event

an Application does not supply the function or data, the facility still operates, and

performs whatever part of its purpose remains meaningful, or

"b) under the GNU GPL, with none of the additional permissions of this License applicable to

that copy.

3. Object Code Incorporating Material from Library Header Files.

The object code form of an Application may incorporate material from a header file that is part of

the Library. You may convey such object code under terms of your choice, provided that, if the

incorporated material is not limited to numerical parameters, data structure layouts and

accessors, or small macros, inline functions and templates (ten or fewer lines in length), you do

both of the following:

a) Give prominent notice with each copy of the object code that the Library is used in it and

that the Library and its use are covered by this License.

b) Accompany the object code with a copy of the GNU GPL and this license document.

112 GNU Lesser General Public License

Page 131 of the Emerson 3812 user manual — B.2 GNU Lesser General Public License
Page 131 · click to enlarge
Page 131

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Open source licenses 3-9000-765 Rev H May 2015

4. Combined Works.

You may convey a Combined Work under terms of your choice that, taken together, effectively

do not restrict modification of the portions of the Library contained in the Combined Work and

reverse engineering for debugging such modifications, if you also do each of the following:

a) Give prominent notice with each copy of the Combined Work that the Library is used in it

and that the Library and its use are covered by this License.

b) Accompany the Combined Work with a copy of the GNU GPL and this license document.

c) For a Combined Work that displays copyright notices during execution, include the

copyright notice for the Library among these notices, as well as a reference directing the

user to the copies of the GNU GPL and this license document.

d) Do one of the following:

0. Convey the Minimal Corresponding Source under the terms of this License, and the Corresponding Application Code in a form suitable for, and under terms that permit, the user to recombine or relink the Application with a modified version of the Linked Version to produce a modified Combined Work, in the manner specified by section 6 of the GNU GPL for conveying Corresponding Source.
1. Use a suitable shared library mechanism for linking with the Library. A suitable mechanism is one that (a) uses at run time a copy of the Library already present on the user's computer system, and (b) will operate properly with a modified version of the Library that is interface-compatible with the Linked Version.

e) Provide Installation Information, but only if you would otherwise be required to provide such information under section 6 of the GNU GPL, and only to the extent that such information is necessary to install and execute a modified version of the Combined Work produced by recombining or relinking the Application with a modified version of the Linked Version. (If you use option 4d0, the Installation Information must accompany the Minimal Corresponding Source and Corresponding Application Code. If you use option 4d1, you must provide the Installation Information in the manner specified by section 6 of the GNU GPL for conveying Corresponding Source.)

GNU Lesser General Public License 113

Page 132 of the Emerson 3812 user manual — B.2 GNU Lesser General Public License
Page 132 · click to enlarge
Page 132

Open source licenses Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

5. Combined Libraries.

You may place library facilities that are a work based on the Library side by side in a single

library together with other library facilities that are not Applications and are not covered by this

License, and convey such a combined library under terms of your choice, if you do both of the

following:

a) Accompany the combined library with a copy of the same work based on the Library,

uncombined with any other library facilities, conveyed under the terms of this License.

b) Give prominent notice with the combined library that part of it is a work based on the

Library, and explaining where to find the accompanying uncombined form of the same

work.

6. Revised Versions of the GNU Lesser General Public License.

The Free Software Foundation may publish revised and/or new versions of the GNU Lesser

General Public License from time to time. Such new versions will be similar in spirit to the

present version, but may differ in detail to address new problems or concerns.

Each version is given a distinguishing version number. If the Library as you received it specifies

that a certain numbered version of the GNU Lesser General Public License "or any later version"

applies to it, you have the option of following the terms and conditions either of that published

version or of any later version published by the Free Software Foundation. If the Library as you

received it does not specify a version number of the GNU Lesser General Public License, you may

choose any version of the GNU Lesser General Public License ever published by the Free

Software Foundation.

If the Library as you received it specifies that a proxy can decide whether future versions of the

GNU Lesser General Public License shall apply, that proxy's public statement of acceptance of

any version is permanent authorization for you to choose that version for the Library.

114 GNU Lesser General Public License

Page 133 of the Emerson 3812 user manual — B.3 BSD Open Source License
Page 133 · click to enlarge
Page 133

Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Open source licenses 3-9000-765 Rev H May 2015

B.3 BSD Open Source License

For more details about the Open SourceTM BSD license or the Open Source Initiative, follow the link below:

http://www.opensource.org/licenses/bsd-license.php

Copyright (c) ,

All rights reserved.

  • Redistribution and use in source and binary forms, with or without modification, are

permitted provided that the following conditions are met:

  • Redistributions of source code must retain the above copyright notice, this list of conditions

and the following disclaimer.

  • Redistributions in binary form must reproduce the above copyright notice, this list of

conditions and the following disclaimer in the documentation and/or other materials provided with the distribution.

  • Neither the name of Daniel® nor the names of its contributors may be used to endorse or

promote products derived from this software without specific prior written permission.

THIS SOFTWARE IS PROVIDED BY THE COPYRIGHT HOLDERS AND CONTRIBUTORS "AS IS" AND

ANY EXPRESS OR IMPLIED WARRANTIES, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED

WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE

DISCLAIMED. IN NO EVENT SHALL THE COPYRIGHT OWNER OR CONTRIBUTORS BE LIABLE FOR

ANY DIRECT, INDIRECT, INCIDENTAL, SPECIAL, EXEMPLARY, OR CONSEQUENTIAL DAMAGES

(INCLUDING, BUT NOT LIMITED TO, PROCUREMENT OF SUBSTITUTE GOODS OR SERVICES;

LOSS OF USE, DATA, OR PROFITS; OR BUSINESS INTERRUPTION) HOWEVER CAUSED AND ON

ANY THEORY OF LIABILITY, WHETHER IN CONTRACT, STRICT LIABILITY, OR TORT (INCLUDING

NEGLIGENCE OR OTHERWISE) ARISING IN ANY WAY OUT OF THE USE OF THIS SOFTWARE,

EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGE.

BSD Open Source License 115

Page 134 of the Emerson 3812 user manual — B.4 M.I.T License
Page 134 · click to enlarge
Page 134

Open source licenses Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

B.4 M.I.T License

For more details about the Open SourceTM MIT license or the Open Source Initiative follow the link below:

http://www.opensource.org/licenses/mit-license.php

Copyright (c)

Permission is hereby granted, free of charge, to any person obtaining a copy of this software and

associated documentation files (the "Software"), to deal in the Software without restriction,

including without limitation the rights to use, copy, modify, merge, publish, distribute,

sublicense, and/or sell copies of the Software, and to permit persons to whom the Software is

furnished to do so, subject to the following conditions:

The above copyright notice and this permission notice shall be included in all copies or

substantial portions of the Software.

THE SOFTWARE IS PROVIDED "AS IS", WITHOUT WARRANTY OF ANY KIND, EXPRESS OR

IMPLIED, INCLUDING BUT NOT LIMITED TO THE WARRANTIES OF MERCHANTABILITY, FITNESS

FOR A PARTICULAR PURPOSE AND NONINFRINGEMENT. IN NO EVENT SHALL THE AUTHORS OR

COPYRIGHT HOLDERS BE LIABLE FOR ANY CLAIM, DAMAGES OR OTHER LIABILITY, WHETHER IN

AN ACTION OF CONTRACT, TORT OR OTHERWISE, ARISING FROM, OUT OF OR IN CONNECTION

WITH THE SOFTWARE OR THE USE OR OTHER DEALINGS IN THE SOFTWARE.

116 M.I.T License

The MIT License

Page 135 of the Emerson 3812 user manual — Table of contents
Page 135 · click to enlarge
Page 135

Table of contents

DanielTM 3812 Liquid Ultrasonic Flow Meter. Installation Manualp.1Contentsp.9Prefacep.3Daniel customer servicep.3Signal words and symbolsp.4Important safety instructionsp.5Noticep.7Warranty and Limitationsp.8List of Tablesp.11List of Figuresp.13Section 1: Introductionp.151.1 Typical Applicationsp.151.2 Features and benefitsp.161.3 Acronyms, abbreviations and definitionsp.171.4 Daniel MeterLink softwarep.191.5 Daniel 3812 Liquid Ultrasonic Flow meter designp.201.6 Meter specificationsp.271.7 Pre-installation considerationsp.311.8 Safetyp.311.9 Daniel 3812 Certifications and Approvalsp.331.10 FCC compliancep.33Section 2: Mechanical installationp.352.1 Meter piping, lifting and mountingp.352.2 Piping recommendationsp.422.3 Meter safety for hoist rings and lifting slingsp.462.3.1 Use of appropriate safety engineered swivel hoist rings in meter end flangesp.472.3.2 Appropriately rated lifting slingsp.542.4 Mounting requirements in heated or cooled pipelinesp.57Section 3: Electrical installationp.593.1 Cable length TTL modep.593.2 Cable length Open Collector modep.593.3 Grounding meter electronics housingp.603.4 Conduit sealsp.613.4.1 Startup for systems using explosion-proof conduitp.623.4.2 Startup for systems that use flame-proof cablep.633.5 Wiring and I/Op.643.5.1 CPU Module labeling and LED indicatorsp.653.6 I/O connectionsp.693.6.1 Frequency/Digital outputsp.703.6.2 Analog input settingsp.743.6.3 Analog output settingsp.743.6.4 Digital Inputp.743.6.5 DHCP server switch settingsp.743.6.6 Configuration protect switch settingsp.753.6.7 External power source connection and fusep.753.7 Security seal installationp.763.7.1 Direct or remote mount transmitter electronics enclosure sealp.763.7.2 Base Enclosure Security Sealsp.783.7.3 Seal the Junction Box remote mount electronics optionp.793.7.4 Bolted band shroud security sealsp.803.7.5 Clamped band shroud security sealsp.813.7.6 Split shroud security sealsp.823.7.7 Latched band shroud security sealsp.833.7.8 Seal conduit portsp.85Section 4: Configurationp.874.1 Daniel MeterLink setupp.874.2 Field Setup Wizardp.884.3 Using AMS Device Manager to configure the meterp.924.4 Using a Field Communicator to configure the meterp.1074.5 Security seals for the meterp.110Appendix A: Engineering drawingsp.111A.1 Daniel 3812 Liquid Ultrasonic Flow Meter drawingsp.111Appendix B: Open source licensesp.117B.1 GNU General Public Licensep.118B.2 GNU Lesser General Public Licensep.129B.3 BSD Open Source Licensep.133B.4 M.I.T Licensep.134Appendix C Indexp.135C.1 Manual Indexp.135
Page 136 of the Emerson 3812 user manual — Appendix C Index
Page 136 · click to enlarge
Page 136

Index Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual May 2015 3-9000-765 Rev H

incorrect sling attachment42 safety precautions using swivel hoist rings34 Split shroud security seals68, 69 swivel hoist rings33 90 degree angle between slings35 Meter specifications13

clamped band shroud6 latched single band shroud6 split shroud6 Source code for executable files99 Startup

startup and configuration45, 73 systems that use flame-proof cable49 T Transducer

analog Input(s)15 analog output(s)16 connectivity15 Digital Input(s)15 frequency output/digital outputs16 power14 temperature14 transducers14 Meter tube considerations

LT-10 operating temperature14 LT-11 operating temperature14 Transmitter Electronics Enclosure chassis ground46 U Ultrasonic type13 W Wiring and connections

bi-directional flow30 piping unidirectional flow29 MMU4 Mounting in heated or cooled pipelines43 O Open source licenses99 Opensource

Ethernet cable to PC communication53 Ethernet communications53 Modbus ASCII50 Modbus ASCII or RTU50 PC to meter serial connection55 PlantWeb® architecture50 RS-485 full duplex50 TCP/IP protocol50 Wiring and I/O50 Numerics 3812 meter accuracy limits14 body and flange pressure rating range (psi)13 communications15 design6 digital, analog, and frequency inputs15 digital, analog, and frequency outputs16 electronic specifications14 Enclosure materials13 flange types14 linearity13 meter performance13, 14 meter type13 Minimum operating pressure14 preinstallation considerations17 repeatability13 specific gravity14 specifications13 temperature base electronic enclosure14 Velocity range13

BSD Open Source License115 GNU General Public License100 GNU Lesser General Public License111 M.I.T License116 P Piping recommendations28

bidirectional flow30 unidirectional flow29 Port A

communications54 Pre-installation considerations17 Public domain99 S Safety17

ATEX Directive 94/9/EC17 FCC compliance19 Seal the meter

Bolted band shroud security seals66 Clamped band shroud security seals67 Latched band shroud security seals69 Split shroud security seal68 Sealing the unit71 Section73 Securing end caps and shroud62 Serial Port A parameters54 Shroud options6

bolted band shroud6

118 Manual Index

Page 137 of the Emerson 3812 user manual — Appendix C Index
Page 137 · click to enlarge
Page 137
Page 137 of the Emerson 3812 user manual — Appendix C Index
Page 138 of the Emerson 3812 user manual — P/N 3-9000-765
Page 138 · click to enlarge
Page 138

P/N 3-9000-765

Daniel Measurement and Control, Inc. 11100 Brittmoore Park Drive Houston, TX 77041 USA

This product is a core component of the PlantWeb digital plant architecture.

T+1 713-467-6000 F+1 713-827-4805 USA Toll Free 1 888-356-9001 Daniel Measurement Services, Inc. T +1 713-827-6314 www.Daniel.com

Europe: Stirling, Scotland, UK T + 44-1786-433400 Middle East: Africa: Dubai, UAE T +971-4-811-8100 Asia Pacific: Singapore T +65-677-8211

© 2015 Daniel Measurement and Control, Inc. All rights reserved

The Emerson logo is a trademark and service mark of Emerson Electric Co. Daniel Measurement and Control, Inc. (Daniel) are Emerson Process Management business units. The Daniel name and logo are trademarks of Daniel Industries, Inc. The Senior, Junior, Simplex, MeterLink SeniorSonic, JuniorSonic, and DanPac are trademarks of Daniel Industries, Inc. All other trademarks are property of their respective owners.

Emerson 3812 — P/N 3-9000-765 diagram (page 138)
Fig. 61

Rev H

Sources

References & links

Pages referenced in this manual — manufacturer, support and downloads.

Ask about the Emerson 3812

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the Emerson 3812

Be the first to ask — other owners or our team will answer.

Original document

Emerson 3812 manual PDF

138 pages·2.6 MB·Oxygen Equipment·PDF
Click any page above to open it here
Page 1 / …
100%
Loading pages…