Installation Manual Par t N u m b er 3 - 90 0 0 - 7 5 9 Revi s i o n D
J u n e 2 0 14 DanielTM 3410 Series Gas Ultrasonic Flow Meters






Return Material Authorization (RMA)
A Return Material Authorization (RMA) number must be obtained prior to returning any equipment for any reason. Download the RMA form from the Support Services web page by selecting the link below.
www2.emersonprocess.com/EN-US/BRANDS/DANIEL/SUPPORT-SERVICES/Pages/Support-Services.aspx?

Pay special attention to the following signal words, safety alert symbols and statements:
This is a safety alert symbol. It is used to alert you to potential physical injury hazards. Obey all safety messages that follow this symbol to avoid possible injury or death.
Safety alert symbol

Daniel Measurement and Control, Inc. (Daniel) designs, manufactures and tests products to function within specific conditions. Because these products are sophisticated technical instruments, it is important that the owner and operation personnel strictly adhere both to the information printed on the product and to all instructions provided in this manual prior to installation, operation, and maintenance.
Daniel also urges you to integrate this manual into your training and safety program.
BE SURE ALL PERSONNEL READ AND FOLLOW THE INSTRUCTIONS IN THIS MANUAL AND ALL NOTICES AND PRODUCT WARNINGS.
Installing, operating or maintaining a Daniel product improperly could lead to serious injury or death from explosion or exposure to dangerous substances. To reduce this risk:
Product owners (Purchasers):

Product Operation Personnel:

THE CONTENTS OF THIS PUBLICATION ARE PRESENTED FOR INFORMATIONAL PURPOSES ONLY, AND WHILE EVERY EFFORT HAS BEEN MADE TO ENSURE THEIR ACCURACY, THEY ARE NOT TO BE CONSTRUED AS WARRANTIES OR GUARANTEES, EXPRESSED OR IMPLIED, REGARDING THE PRODUCTS OR SERVICES DESCRIBED HEREIN OR THEIR USE OR APPLICABILITY. ALL SALES ARE GOVERNED BY DANIEL'S TERMS AND CONDITIONS, WHICH ARE AVAILABLE UPON REQUEST. WE RESERVE THE RIGHT TO MODIFY OR IMPROVE THE DESIGNS OR SPECIFICATIONS OF SUCH PRODUCTS AT ANY TIME.
DANIEL DOES NOT ASSUME RESPONSIBILITY FOR THE SELECTION, USE OR MAINTENANCE OF ANY PRODUCT. RESPONSIBILITY FOR PROPER SELECTION, USE AND MAINTENANCE OF ANY DANIEL PRODUCT REMAINS SOLELY WITH THE PURCHASER AND END-USER.
TO THE BEST OF DANIEL'S KNOWLEDGE THE INFORMATION HEREIN IS COMPLETE AND ACCURATE. DANIEL MAKES NO WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING THE IMPLIED WARRANTIES OF MERCHANTABIL- ITY AND FITNESS FOR A PARTICULAR PURPOSE WITH RESPECT TO THIS MANUAL AND, IN NO EVENT, SHALL DANIEL BE LIABLE FOR ANY INCIDENTAL, PUNITIVE, SPECIAL OR CONSEQUENTIAL DAMAGES INCLUDING, BUT NOT LIMITED TO, LOSS OF PRODUCTION, LOSS OF PROFITS, LOSS OF REVENUE OR USE AND COSTS INCURRED INCLUDING WITHOUT LIMITATION FOR CAPITAL, FUEL AND POWER, AND CLAIMS OF THIRD PARTIES.
PRODUCT NAMES USED HEREIN ARE FOR MANUFACTURER OR SUPPLIER IDENTIFICATION ONLY AND MAY BE TRADEMARKS/REGISTERED TRADEMARKS OF THESE COMPANIES

1. LIMITED WARRANTY: Subject to the limitations contained in Section 2 herein, Daniel Measurement & Control, Inc. (“Daniel”) warrants that the licensed firmware embodied in the Goods will execute the programming instructions provided by Daniel, and that the Goods manufactured by Daniel will be free from defects in materials or workmanship under normal use and care and Services will be performed by trained personnel using proper equipment and instrumentation for the particular Service provided. The foregoing warranties will apply until the expiration of the applicable warranty period. Goods are warranted for twelve (12) months from the date of initial installation or eighteen (18) months from the date of shipment by Daniel, whichever period expires first. Consumables and Services are warranted for a period of 90 days from the date of shipment or completion of the Services. Products purchased by Daniel from a third party for resale to Buyer ("Resale Products”) shall carry only the warranty extended by the original manufacturer. Buyer agrees that Daniel has no liability for Resale Products beyond making a reasonable commercial effort to arrange for procurement and shipping of the Resale Products. If Buyer discovers any warranty defects and notifies Daniel thereof in writing during the applicable warranty period, Daniel shall, at its option, correct any errors that are found by Daniel in the firmware or Services or repair or replace F.O.B. point of manufacture that portion of the Goods or firmware found by Daniel to be defective, or refund the purchase price of the defective portion of the Goods/Services. All replacements or repairs necessitated by inadequate maintenance, normal wear and usage, unsuitable power sources or environmental conditions, accident, misuse, improper installation, modification, repair, use of unauthorized replacement parts, storage or handling, or any other cause not the fault of Daniel are not covered by this limited warranty, and shall be at Buyer's expense. Daniel shall not be obligated to pay any costs or charges incurred by Buyer or any other party except as may be agreed upon in writing in advance by Daniel. All costs of dismantling, reinstallation and freight and the time and expenses of Daniel's personnel and representatives for site travel and diagnosis under this warranty clause shall be borne by Buyer unless accepted in writing by Daniel. Goods repaired and parts replaced by Daniel during the warranty period shall be in warranty for the remainder of the original warranty period or ninety (90) days, whichever is longer. This limited warranty is the only warranty made by Daniel and can be amended only in a writing signed by Daniel. THE WARRANTIES AND REMEDIES SET FORTH ABOVE ARE EXCLUSIVE. THERE ARE NO REPRESENTATIONS OR WARRANTIES OF ANY KIND, EXPRESS OR IMPLIED, AS TO MERCHANTABILITY, FITNESS FOR PARTICULAR PURPOSE OR ANY OTHER MATTER WITH RESPECT TO ANY OF THE GOODS OR SERVICES. Buyer acknowledges and agrees that corrosion or erosion of materials is not covered by this warranty.
2. LIMITATION OF REMEDY AND LIABILITY: DANIEL SHALL NOT BE LIABLE FOR DAMAGES CAUSED BY DELAY IN PERFORMANCE. THE REMEDIES OF BUYER SET FORTH IN THIS AGREEMENT ARE EXCLUSIVE. IN NO EVENT, REGARDLESS OF THE FORM OF THE CLAIM OR CAUSE OF ACTION (WHETHER BASED IN CONTRACT, INFRINGEMENT, NEGLIGENCE, STRICT LIABILITY, OTHER TORT OR OTHERWISE), SHALL DANIEL'S LIABILITY TO BUYER AND/OR ITS CUSTOMERS EXCEED THE PRICE TO BUYER OF THE SPECIFIC GOODS MANUFACTURED OR SERVICES PROVIDED BY DANIEL GIVING RISE TO THE CLAIM OR CAUSE OF ACTION. BUYER AGREES THAT IN NO EVENT SHALL DANIEL'S LIABILITY TO BUYER AND/OR ITS CUSTOMERS EXTEND TO INCLUDE INCIDENTAL, CONSEQUENTIAL OR PUNITIVE DAMAGES. THE TERM “CONSEQUENTIAL DAMAGES” SHALL INCLUDE, BUT NOT BE LIMITED TO, LOSS OF ANTICIPATED PROFITS, REVENUE OR USE AND COSTS INCURRED INCLUDING WITHOUT LIMITATION FOR CAPITAL, FUEL AND POWER, AND CLAIMS OF BUYER'S CUSTOMERS.


Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual List of Tables 3-9000-759 Rev D June 2014
Table 1-1 Daniel 3410 Series Gas Ultrasonic Meter acronyms, abbreviations and definitions 3
Table 1-3 Performance specifications 13
Table 1-2 Meter specifications 13
Table 1-4 Transducer specifications 15
Table 2-1 Hoist ring part number lookup table 38
Table 2-2 Hoist Ring Lookup Table for Daniel 3414 Gas Meters 39
Table 2-3 Hoist Ring Lookup Table for Daniel 3411 or 3412 Gas Meters 39
Table 3-1 Configurations for open collector frequency outputs 43
Table 3-2 CPU Module labeling and LED functions 53
Table 3-3 Ethernet cable to PC communication 54
Table 3-4 Serial Port A parameters 56
Table 3-5 Frequency/Digital Outputs possible configurations 61
Table 3-6 DHCP server switch settings 63
Table 3-7 Configuration protect switch settings 64
Table 4-1 Local display labels, descriptions and valid units 77
Table B-1 Open source licences 101
List of Tables v

List of Tables Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
vi List of Tables

Daniel 3410 Series Gas Ultrasonic Meter Installation Manual List of Figures 3-9000-759 Rev D June 2014
Figure 1-1 Daniel MeterLink download and registration 6
Figure 1-2 Daniel 3414 Gas Ultrasonic Flow Meter design 8
Figure 1-3 Daniel 3412 Gas Ultrasonic Flow Meter design 9
Figure 1-4 Daniel 3411 Gas Ultrasonic Flow Meter design 10
Figure 1-5 Transmitter electronics enclosure with local display and glass endcap 10
Figure 2-1 Daniel 3414 Flow Meter assembly 26
Figure 2-2 Daniel 3412 Flow Meter assembly 27
Figure 2-3 Daniel 3411 Flow Meter assembly 28
Figure 2-4 Transmitter electronics enclosure with optional local display and glass endcap 28
Figure 2-5 Piping recommendations uni-directional without flow conditioner 30
Figure 2-6 Piping recommendations Uni-directional with flow conditioner 31
Figure 2-7 Piping recommendations Bi-directional flow with flow conditioner 31
Figure 2-8 Meter end flange with tapped flat-counterbore hole for hoist ring 34
Figure 2-9 Safety approved hoist ring and non-compliant eye bolt 35
Figure 2-10 90 Degree angle between slings 36
Figure 2-11 Incorrect sling attachment 37
Figure 2-12 Correct sling attachment 40
Figure 2-13 Incorrect sling attachment 41
Figure 3-1 Internal Transmitter Electronics Enclosure chassis ground 44
Figure 3-2 External ground lug 45
Figure 3-3 Electronics field wiring - upper terminal block, switches, ground lug 47
Figure 3-4 Transmitter electronics field wiring lower terminal block 48
Figure 3-5 CPU Module labeling and LED indicators 52
Figure 3-6 PC to meter serial connection wiring 57
Figure 3-7 CPU Module I/O connections 58
Figure 3-8 CPU Module - Frequency/Digital outputs common ground 62
Figure 3-9 CPU Module power source connections 65
Figure 3-10 Transmitter electronics enclosure security latch 66
Figure 3-11 Transmitter Electronics Enclosure security seals 67
List of Figures vii

List of Figures Daniel 3410 Series Gas Ultrasonic Meter Installation Manual June 2014 3-9000-759 Rev D
Figure 3-12 Base Enclosure wire seal installation 68
Figure 3-13 Base Enclosure security seals 69
Figure 3-14 Transducer assembly security seal70
Figure 4-1 AMS Device Description search81
Figure 4-2 AMS file download complete 82
Figure 4-3 AMS Device Manager 83
Figure 4-4 AMS Device Manager - Overview 83
Figure 4-5 AMS Device Manager - Guided Setup 84
Figure 4-6 AMS Device Manager - Service Tools All Variables status indicators86
Figure 4-7 Display Meter K-Factors86
Figure 4-8 AMS Device Manager - Configure Manual Setup87
Figure 4-9 Gating configuration parameter Edge gated, active high 89
Figure 4-10 Gating configuration parameter Edge gated, active low89
Figure 4-11 Gating configuration parameter State gated, active high89
Figure 4-12 Gating configuration parameter State gated, active low89
Figure 4-13 Configure Flow Analysis Alert 90
Figure 4-14 AMS Device Manager - Service Tools Alerts 91
Figure 4-15 Configuration changes dialog91
Figure 4-16 AMS Device Manager - Service Tools 92
Figure 4-17 AMS Device Manager - Service Tools All Variables 93
Figure 4-18 AMS Device Manager - Service Tools Trends 94
Figure 4-19 3414 transmitter field wiring conduit entries 96
Figure 4-20 Field Communicator wiring diagram for the 3410 Series electronics 97
viii List of Figures

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 1: Introduction
3-9000-759 Rev D June 2014
Daniel 3410 Series Ultrasonic Gas Flow Meters have various configurations that meet a broad range of customer requirements. Each meter comes fully assembled from DanielTM Measurement and Control, Inc.
Refer to the following documents for additional details:
Daniel 3414 Gas Ultrasonic Flow Meter technology provides custody transfer gas measurement with the 3414 Model four-path (eight transducers) chordal design.
Model 3412 dual-path (four transducers) chordal design and Model 3411 single-path (two transducers) chordal design provides check metering, pipeline balance, storage measurement, production, or wet gas applications.
See Section 1.2 for advantages of Daniel 3410 Series Ultrasonic Gas Flow Meters.
Daniel 3410 Series Ultrasonic Gas Flow Meters have various configurations that meet a broad range of customer requirements. Each meter comes fully assembled from Daniel. The technology can be applied to custody transfer, allocation measurement, and check metering applications such as:
Typical applications 1

June 2014 3-9000-759 Rev D
— Abnormal profile — Blockage — Internal bore buildup — Liquids present in the gas meter — Reverse Flow — Speed of Sound comparison error
For other features and benefits refer to the product datasheet:
http://www2.emersonprocess.com/EN-US/BRANDS/DANIEL/FLOW/Pages/Flow.aspx
2 Features and benefits

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 1: Introduction
3-9000-759 Rev D June 2014
Table 1-1 Daniel 3410 Series Gas Ultrasonic Meter acronyms, abbreviations and definitions
Acronym or abbreviation Definition
oC degrees celsius (temperature unit)
oF degrees fahrenheit (temperature unit)
AMS® Device Manager Asset Management Software - Device Manager
ASCII MODBUS A Modbus protocol message framing format in which ASCII characters are used to delineate the beginning and end of the frame. ASCII stands for American Standard Code for Information Interchange.
boolean A type of data point that can only take on values of TRUE or FALSE (generally TRUE is represented by a value of 1, FALSE is represented by a value of 0)
bps Bits Per Second (baud rate)
cPoise centipoise (viscosity unit)
CPU Central Processing Unit
CTS Clear-to-Send; the RS-232C handshaking signal input to a transmitter indicating that it is okay to transmit data – i.e., the corresponding receiver is ready to receive data. Generally, the Request-to-Send (RTS) output from a receiver is input to the Clear-to- Send (CTS) input of a transmitter.
Daniel MeterLinkTM Daniel Ultrasonic Meter interface software
DHCP Dynamic Host Configuration Protocol
dm decimeter (10-1 meters, length unit)
ECC Error Correction Code
EEPROM Electrically-Erasable, Programmable Read-Only Memory
Flash non-volatile, programmable read-only memory
FODO output that is user configurable as either a Frequency or Digital Output
HART® Communication Protocol Highway Addressable Remote Transducer communications protocol
hr hour (time unit)
Hz Hertz (cycles per second, frequency unit)
Acronyms, abbreviations and definitions 3

June 2014 3-9000-759 Rev D
Table 1-1 Daniel 3410 Series Gas Ultrasonic Meter acronyms, abbreviations and definitions
Acronym or abbreviation Definition
K Kelvin (temperature unit)
kHz kilohertz (103 cycles per second, frequency unit)
LAN Local Area Network
m meter (length unit)
m3/d cubic meters per day (volumetric flow rate)
m3/h cubic meters per hour (volumetric flow rate)
m3/s cubic meters per second (volumetric flow rate)
mA milliamp (current unit)
MAC Address Media Access Control (Ethernet Hardware Address -EHA)
microinch (μinch) microinch (10-6 in)
micron micrometer (10-6 m)
MMU Memory Management Unit
MPa megapascal (equivalent to 106 Pascal) (pressure unit)
Nm3/h normal cubic meters per hour
NOVRAM Non-Volatile Random Access Memory
Pa Pascal, equivalent to 1 newton per square meter (pressure unit)
Pa⋅s Pascal Second (viscosity unit)
PFC Peripheral Field Connection (Board)
PS Power Supply (board)
psi pounds per square inch (pressure unit)
psia pounds per square inch absolute (pressure unit)
psig pounds per square inch gage (pressure unit)
R Radius of meter
RAM Random Access Memory
RTS Request-to-Send; the RS-232C handshaking signal output by a receiver when it is ready to receive data
RTU MODBUS A Modbus protocol framing format in which elapsed time between received charac- ters is used to separate messages. RTU stands for Remote Terminal Unit.
4 Acronyms, abbreviations and definitions

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 1: Introduction
3-9000-759 Rev D June 2014
Table 1-1 Daniel 3410 Series Gas Ultrasonic Meter acronyms, abbreviations and definitions
Acronym or abbreviation Definition
s second (time unit, metric)
SDRAM Synchronous Dynamic Random Access Memory
sec second (time unit, U.S. Customary)
TCP/IP Transmission Control Protocol/Internet Protocol
time_t seconds since Epoch (00:00:00 UTC Jan. 1, 1970) (time unit)
UDP User Datagram Protocol
U.L. Underwriters Laboratories, Inc. - product safety testing and certification organization
V Volts (electric potential unit)
W Watts (power unit)
Acronyms, abbreviations and definitions 5

June 2014 3-9000-759 Rev D
Daniel MeterLinkTM software has robust features for setting communications parameters, configuring your meter, collecting logs and reports and monitoring the meter health and alarm statuses. Daniel MeterLink may be downloaded at no charge from:
Figure 1-1 Daniel MeterLink download and registration

6 Daniel MeterLink software

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 1: Introduction
3-9000-759 Rev D June 2014
Refer to the Daniel MeterLink Software for Gas and Liquid Ultrasonic Meters Quick Start Manual (P/N 3-9000-763) for installation instructions and setup for initial communications. You may download the manual from the Daniel MeterLink web page:
http://www2.emersonprocess.com/en-US/brands/daniel/Flow/ultrasonics/Pages/MeterLink.aspx
Daniel 3410 Series Gas Ultrasonic Flow Meters are designed to accurately measure products in applications where reliable performance is critical, by measuring the difference in signal transit time with and against the flow across one or more measurement path(s). A signal transmitted in the flow direction travels faster than one transmitted against the flow direction. Each measurement path is defined by a transducer pair in which each transducer alternately acts as transmitter and receiver. The meter uses transit time measurements and transducer location information to calculate the mean velocity.
Daniel 3410 Series meter design 7

June 2014 3-9000-759 Rev D
Computer simulations of various velocity profiles demonstrate that multiple measurement paths provide an optimum solution for measuring asymmetric flow.Daniel 3414 Gas Ultrasonic Flow meters utilize four cross-bore, parallel-plane measurement paths that offer a high degree of accuracy, repeatability, bi-directional measurement and superior low-flow capabilities without the compromises associated with conventional technologies. These features make the Daniel 3414 the best choice for custody transfer applications.
Figure 1-2 Daniel 3414 Gas Ultrasonic Flow Meter design
A.
B.
A. Transmitter electronics enclosure (explosion-proof) Optional - Local Display with glass endcap (Figure 1-4) B. Base electronics enclosure (intrinsically safe) C. Meter body with transducer assemblies (T-11, T-12, T-21 or T-22) (intrinsically safe)
8 Daniel 3410 Series meter design

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 1: Introduction
3-9000-759 Rev D June 2014
Daniel 3412 Gas Ultrasonic Flow meters utilize two-path in-line (four transducers) measurement paths and are designed to measure the difference in signal transit time with and against the flow across one or more measurement path(s). The two paths are configured at right angles to one another in a “bulls-eye” arrangement.
Figure 1-3 Daniel 3412 Gas Ultrasonic Flow Meter design
A.
B.
A. Transmitter electronics enclosure (explosion-proof) Optional - Local Display with glass endcap (Figure 1-4) B. Base electronics enclosure (intrinsically safe) C. Meter body with transducer assemblies (T-11, T-12, T-21 or T-22) (intrinsically safe)
Daniel 3410 Series meter design 9

June 2014 3-9000-759 Rev D
Daniel 3411 Gas Ultrasonic Flow meters are single-path (two transducer) Gas Ultrasonic Flow Meter and is referred to as a bounce-path (as the signal is bounced off the meter body) or a centerline path (as it goes through the centerline of the meter body) meter. The bounce-path method simplifies construction of the meter and makes the meter less susceptible to interference from pipeline liquids.
Figure 1-4 Daniel 3411 Gas Ultrasonic Flow Meter design

A.
B.
A. Transmitter electronics enclosure (explosion-proof) Optional - Local Display with glass endcap (Figure 1-4) B. Base electronics enclosure (intrinsically safe) C. Meter body with transducer assemblies (T-11, T-12, T-21 or T-22) (intrinsically safe)
The Daniel Gas Ultrasonic Flow Meter design is available with an optional glass endcap and a local display.
Figure 1-5 Transmitter electronics enclosure with local display and glass endcap

10 Daniel 3410 Series meter design

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 1: Introduction
3-9000-759 Rev D June 2014
All Daniel ultrasonic flow meter’s U.L. safety listing is accomplished through the combination of an explosion-proof transmitter electronics enclosure that houses the CPU module, Power Supply board, I.S. Barrier board, Backplane board and optional LCD Display board.
The Base Electronics Enclosure that houses the Acquisition Module. Intrinsically safe transducers and cable assemblies are designed for Class 1, Division 1, Groups C and D areas without need of further protection when installed in accordance with the field wiring diagram (refer to Daniel drawing DMC - 005324 (see Appendix A).
Daniel 3410 Series meter design 11

June 2014 3-9000-759 Rev D
When the meter is under pressure, DO NOT attempt to remove or adjust the transducer holder.
Attempting to do so may release pressurized gases, resulting in serious injury or equipment damage
The meter must be fully depressurized and drained before attempting to remove the transducer holder. If gas or fluid begins to leak from the transducer holder, stop immediately and reinstall the holder.
Failure to do so may cause serious injury or equipment damage.

A.
A. Transducer holder
The purchaser of the meter is responsible for the selection of Daniel components/seals and materials compatible with the chemical properties of gas flow measurement.
Failure to select suitable meter components/seals may cause escaping gases or liquids, resulting in injury or equipment damage.
Consult your Daniel Sales and Service representative to ensure you purchase the correct components and seals for your application.
12 Meter specifications

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 1: Introduction
3-9000-759 Rev D June 2014
Specifications for Daniel 3410 Series Gas Ultrasonic Flow Meters are below:
Table 1-2 Meter specifications
Daniel 3411, 3412 and 3414 meter specifications
— 100% conversion coated and exterior coated with a polyurethane enamel • ASTM A351 Gr CF8M Stainless Steel
— Passivated Optional Local Display with a glass endcap on transmitter enclosure
Linearity • Model 3414 Four-path meter chordal design
— ± 0.3% of measured value over a 100:1 turndown 3-100 ft/s; 0.3 to 30 m/s) including lab uncertainty — Flow calibrated accuracy is ± 0.1% of reading relative to lab over entire flow calibration range (Qmin - Qmax) • Model(s) 3411 single-path or 3412 two-path
— Flow calibrated accuracy is ± 0.5% of reading relative to lab1
— Accuracy is typically ±1.5% of actual volume flow1(without flow calibration) 1. Does not take into consideration changes in wall roughness and installation effects. Repeatability • ±0.05% of reading in the specified velocity range from 5% to 100% (Qmax)
Table 1-3 Performance specifications
Meter size 4” to 24” 30” 36”
Qmin(ft/s) 2 2 2
Qt (ft/s) 10 8.5 7.5
Qmax (ft/s) 100 85 75
Body and Flange Sizes and Pressure rating range
Meter Performance · Meter specifications 13

June 2014 3-9000-759 Rev D
Table 1-2 Meter specifications
100 psig (7 bar)
Meter Performance · Electronic specifications · 14 Meter specifications

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 1: Introduction
3-9000-759 Rev D June 2014
Table 1-2 Meter specifications
Transducer, mounts and holders
Table 1-4 Transducer specifications
Transducer type Temperature range Mount and holder type
T-11 -20 °C to +100 °C (-4 °F to 212 °F) Standard mounts/Holders, NBR O-ring Inconel mounts/316L Holders, NBR O-ring Inconel Mounts/Inconel Holders/FKM O-ring
T-12 -20 °C to +100 °C (-4 °F to 212 °F) Standard mounts/Holders, NBR O-ring Inconel mounts/316L Holders, NBR O-ring Inconel Mounts/Inconel Holders/FKM O-ring
T-211 -20 °C to +100 °C (-4 °F to 212 °F) Standard mounts/Holders, NBR O-ring Inconel mounts/316L Holders, NBR O-ring Inconel Mounts/Inconel Holders/FKM O-ring
T-222 -50 °C to +100 °C (-58°F to 212 °F) Standard mounts/Holders, NBR O-ring Inconel mounts/316L Holders, NBR O-ring Inconel Mounts/Inconel Holders/FKM O-ring
through 10 meters.
transmitter electronics cannot be remote mounted from Division 1 classification to a Division 2 area to meet an area classification.
Electronic specifications (Continued) · Meter specifications 15

June 2014 3-9000-759 Rev D
Table 1-2 Meter specifications
Digital, analog, and frequency inputs
temperature range.
resistance so HART® Communicators can be connected to configure sensors.
A 24 Volt DC power output is available to provide power to the sensors.

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 1: Introduction
3-9000-759 Rev D June 2014
Table 1-2 Meter specifications
Digital, analog, and frequency outputs
The analog output zero scale offset error is within ±0.1% of full scale and gain error is within ±0.2% of full scale. The total output drift is within ±50 ppm of full scale per °C.
Frequency/Digital Output(s) · Meter specifications 17

June 2014 3-9000-759 Rev D
The Daniel 3410 Series Gas Ultrasonic Flow Meter is suitable for use in U.L. Class 1, Division 1, Group C and D hazardous locations.
An “X” signifies the user should contact Daniel Measurement and Control, Inc. for information on the dimensions of the flameproof joints.
Refer to the 3410 Series Systems Wiring Diagram, Sheet 3 (P/N DMC -005324) for the certification tag (see Appendix A).
18 Pre-installation considerations

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 1: Introduction
3-9000-759 Rev D June 2014
Daniel 3410 Series Liquid Ultrasonic Meters are INMETRO certified. Refer to the 3410 Series Liquid Ultrasonic Flow Meter Tag, INMETRO Certification drawing DMC - 006224.
Certificate number: NCC 11.0163 X Marking: --Ex d ia IIB T4 Gb IP66 W Electrical parameters: Refer to Section 1.6, Table 1-1
Special conditions for safe use
40 oF (60 oC) (refer to Section 1.6, Table 1-2)
Conduit runs must have a sealing fitting within 18 inches (457 mm) of the enclosure to reduce the risk of an explosion or a fire.
Failure to do so may result in severe injury to personnel or cause damage to the equipment.
Safety 19

June 2014 3-9000-759 Rev D
Daniel 3410 Series Gas Ultrasonic Flow Meters have electrical, metrology, intrinsic safety and Pressure Equipment Directive certifications and approvals by the agencies listed below. Refer to the nameplate tag on the meter body, the wiring diagram (P/N DMC - 005324) in Appendix A and observe all safety precautions. Daniel 3410 Series Gas Ultrasonic Flow Meters operate within the pressure and temperature range of the device (also see Section 1.6 for meter specifi- cations). Daniel 3410 Series Gas Ultrasonic Flow Meters are approved to the ATEX Directive 94/9/EC.
- Explosive Atmospheres (ATEX) - International Electrotechnical Commission (IECEx) - Pressure Equipment Directive (PED via BSI) - Electromagnetic Compatibility (EMC) - International Organization of Legal Metrology (OIML)
20 Daniel 3410 Series Certifications and Approvals

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 1: Introduction
3-9000-759 Rev D June 2014
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment.
This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
Changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment.
FCC compliance 21

June 2014 3-9000-759 Rev D
[1] Gould Modbus Protocol Reference Guide, Rev. B, PI-MBUS-300
[2] Measurement of Fuel Gas By Turbine Meters, American Gas Association, Transmission
Measurement Committee Report No. 7, Second Revision, April 1996 (also referred to as AGA7)
[3] Compressibility Factors of Natural Gas and Other Related Hydrocarbon Gases, American
Gas Association, Transmission Measurement Committee Report No. 8, Second Edition, Second Printing, July 1994 (also referred to as AGA8)
[4] Speed of Sound in Natural Gas and Other Related Hydrocarbon Gasses, Report 10, First
Edition, May 2003 (also referred to as AGA10)
[5] Manual of Petroleum Measurement Standards, Chapter 21 — Flow Measurement Using
Electronic Metering Systems, Section 1 — Electronic Gas Measurement, American Gas Association and American Petroleum Institute, First Edition, September 1993
[6] AGA Report No. 9, Measurement of Gas by Multipath Ultrasonic Meters, Second Edition
(April 2007) · 22 References

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 2: Mechanical installation 3-9000-759 Rev D June 2014
Refer to the following sections for piping recommendations, lifting with hoist rings and slings, mounting in heated or cooled pipelines and safety warnings and precautions.
The meter body and piping may be extremely hot or cold.
Wear appropriate personal protective equipment when coming in contact with the meter. Failure to do so may result in injury.
Sharp edges may be present on the meter.
Wear appropriate personal protective equipment when working on the meter. Failure to do so may cause serious injury.
When moving the meter, do not insert the forks of a forklift into the bore.
inserting the forks may cause the meter to become unstable, resulting in injury or damage to the bore and sealing face.
Clear all obstacles or obstructions from the work area when transporting, installing or removing the meter.
Failure to clear the work area may cause injury to personnel.
Meter piping, lifting and mounting 23

Section 2: Mechanical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
Do not remove flange stabilizers.
Attempting to do so may allow the meter to roll, resulting in serious injury or equipment damage.
A.
A. Flange stabilizers
The purchaser of the meter is responsible for the selection of Daniel components/seals and materials compatible with the chemical properties of gas flow measurement.
Failure to select suitable meter components/seals may cause escaping gases or liquids, resulting in injury or equipment damage.
Consult your Daniel Sales and Service representative to ensure you purchase the correct components and seals for your application.
24 Meter piping, lifting and mounting

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 2: Mechanical installation 3-9000-759 Rev D June 2014
Daniel 3410 Series Gas Ultrasonic Flow Meters are assembled, configured and tested at the factory. The meter components include the transmitter electronics enclosure, the base electronics enclosure and the meter body with transducer assemblies. 1
When the meter is under pressure, DO NOT attempt to remove or adjust the transducer holder.
Attempting to do so may release pressurized gases, resulting in serious injury or equipment damage
The meter must be fully depressurized and drained before attempting to remove the transducer holder. If gas or fluid begins to leak from the transducer holder, stop immediately and reinstall the holder.
Failure to do so may cause serious injury or equipment damage.
A.
A. Transducer holder
Meter components 25

Section 2: Mechanical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
Conduit runs must have a sealing fitting within 18 inches (457 mm) of the enclosure to reduce the risk of an explosion or a fire.
Failure to do so may result in severe injury to personnel or cause damage to the equipment.
The 3414 four path ultrasonic meter components are shown below.
Figure 2-1 Daniel 3414 Flow Meter assembly
A.
B.
D.
A. Explosion-proof transmitter enclosure (CPU Module, Power Supply, I.S. Barrier Board
B. Intrinsically-safe base enclosure includes Acquisition Module C. Meter - body and transducer assemblies and cables
D. Flange stabilizers
26 Meter components

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 2: Mechanical installation 3-9000-759 Rev D June 2014
The 3412 dual path ultrasonic meter components are shown below. Figure 2-2 Daniel 3412 Flow Meter assembly
A.
B.
D.
A. Explosion-proof transmitter enclosure (CPU Module, Power Supply, I.S. Barrier Board
B. Intrinsically-safe base enclosure includes Acquisition Module C. Meter - body and transducer assemblies and cables D. Flange stabilizers
Meter components 27

Section 2: Mechanical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
The 3411 single path ultrasonic meter components are shown below.
Figure 2-3 Daniel 3411 Flow Meter assembly
A.
B.
D.
A. Explosion-proof transmitter enclosure (CPU Module, Power Supply, I.S. Barrier Board
B. Intrinsically-safe base enclosure includes Acquisition Module C. Meter - body and transducer assemblies and cables D. Flange stabilizers
Figure 2-4 Transmitter electronics enclosure with optional local display and glass endcap
A.
B.
A. Transmitter electronics enclosure with glass endcap B. Local display
28 Meter components

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 2: Mechanical installation 3-9000-759 Rev D June 2014
Before pipeline cleaning and maintenance ("pigging operations"), remove straightening vanes or flow conditioners. Failure to do so may cause excessive pressure in the meter system, resulting in serious injury/ death or equipment damage.
3410 Series Gas Ultrasonic Flow Meter with flow conditioner for uni-directional flow

3410 Series Gas Ultrasonic Flow Meter with flow conditioner for bi-directional flow

Sunshields, provided by the customer, may be required to prevent exceeding the process fluid temperature when the meter is mounted in a location with extremely hot climates.
Install a sunshield to prevent prolonged exposure to direct sunlight in extreme climates.
Failure to shield the meter may result in exceeding the process temperature range and damage transmitter electronics.
Piping recommendations 29

Section 2: Mechanical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
For optimal flow measurement conditions, Daniel suggests the piping configurations below. Regardless of the configuration selected, the user agrees to accept full responsibility for the site piping design and installation.
Flow conditioning is recommended for best measurement results
The bore of the mating piping should be within 1% of the meter inside diameter.
Figure 2-5 Piping recommendations uni-directional without flow conditioner

30 Piping recommendations

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 2: Mechanical installation 3-9000-759 Rev D June 2014
Figure 2-6 Piping recommendations Uni-directional with flow conditioner

Figure 2-7 Piping recommendations Bi-directional flow with flow conditioner

All pipe lengths are minimum:
http://www2.emersonprocess.com/en-US/brands/daniel/Flow/ultrasonics/Pages/Ultrasonic- Series-3400.aspx
To access the product datasheet from the Daniel products page (above link), select the Daniel Gas Ultrasonic Flow Meter link, click the Documentation tab, expand the Data Sheets - Bulletins - Catalogs tab, then select the Data Sheet.
Piping recommendations 31

Section 2: Mechanical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
Correctly install the equipment.
If meter bodies are mounted or oriented differently than specified above, debris or gas may collect in the transducer ports which could adversely affect the transducer signals, or cause equipment damage.
32 Piping recommendations

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 2: Mechanical installation 3-9000-759 Rev D June 2014
A Daniel Gas Ultrasonic Flow Meter can be safely lifted and maneuvered into and out of a meter run for installation or service by obeying the following instructions.
Lifting a Daniel Ultrasonic Meter with other equipment
The following lifting instructions are for installation and removal of the Daniel Ultrasonic Meter ONLY. The instructions below do not address lifting the Daniel ultrasonic meter while it is attached, bolted, or welded to meter tubes, piping, or other fittings.
Using these instructions to maneuver the Daniel Ultrasonic Meter while it is still attached, bolted, or welded to a meter tube, piping, or other fitting may result in equipment damage, serious injury, or death.
The operator must refer to their company's hoisting and rigging standards, or the "DOE-STD- 1090-2004 Hoisting and Rigging" standard if such company standards do not exist, for lifting and maneuvering any assembled meter tube and associated piping.
During meter installation or removal, always place the unit on a stable platform or surface that supports its assembled weight.
Failure to do so could allow the meter to roll, resulting in serious injury or equipment damage.
Prior to lifting the unit, refer to the Daniel 3414, 3412, or 3411 Gas Ultrasonic Flow Meter nameplate or outline dimensional (general arrangement) drawing for the assembled weight.
Meter safety for hoist rings and lifting slings 33

Section 2: Mechanical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
When lifting a Daniel Ultrasonic Meter by itself, Daniel recommends two methods. These methods are:
Both methods must be used in conjunction with all appropriate company hoisting and rigging standards or the DOE-STD-1090-2004 HOISTING AND RIGGING standard if such company standards do not exist. Refer to the following sections for more information on these two methods.
2.4.1 Use of appropriate safety engineered swivel hoist rings in meter end flanges
Daniel Ultrasonic meters come equipped with a tapped hole located on the top of each meter body end flange. A flat machined surface surrounds each tapped hole. This feature provides complete surface contact ONLY between the meter flange and an OSHA compliant Safety Engineered Swivel Hoist Ring as shown in Figure 2-9.
Operators SHALL NOT use Eye Bolts (see Figure 2-9) in the Daniel Ultrasonic Meter flange tapped holes to aid in lifting or maneuvering the unit.
Operators SHALL NOT use other Hoist Rings that do not fully seat flush with the counter bore on the top of the meter flanges.
Figure 2-8 Meter end flange with tapped flat-counterbore hole for hoist ring

A.
B.
A. Plug Bolt
B. Flat Counterbore Surface
34 Use of appropriate safety engineered swivel hoist rings in meter end flanges

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 2: Mechanical installation 3-9000-759 Rev D June 2014
Figure 2-9 Safety approved hoist ring and non-compliant eye bolt


Eye bolt Safety engineered swivel hoist ring
Read and follow the Safety Precautions listed below:
discard the bolts as they must be reinstalled once the lifting operation is complete to prevent corrosion of the tapped holes.
the hoist rings.
not use any other type of hoist rings with the same screw size or heavy duty hoist rings. The meter tapping and counter bore size are suitable only for the hoist rings specified by Daniel.
the machined flat surface of the tapped hole. If the two surfaces do not come in contact then the hoist ring will not hold its full rated load. Torque the hoist ring attachment bolts to the limit indicated on the hoist rings.
Use of appropriate safety engineered swivel hoist rings in meter end flanges 35

Section 2: Mechanical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
in all directions.
rings. The slings must be of equal length. Each sling must have a load rating that equals or exceeds the hoist ring load rating. The angle between the two slings going to the hoist rings must never exceed 90 degrees or the load rating of the hoist rings will be exceeded.
Figure 2-10 90 Degree angle between slings

36 Use of appropriate safety engineered swivel hoist rings in meter end flanges

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 2: Mechanical installation 3-9000-759 Rev D June 2014
may occur. Use a spreader bar with the slings to prevent contact with the electronics enclosure and the base enclosure (see Figure 2-12). If the slings do come in contact with the electronic enclosure then remove the two bolts holding the enclosure to its base and temporarily remove the head from the meter during the lifting operation. You will need to unplug the cable from J3 on the Acquisition Module. Two screws hold this cable in place.
Once the lifting operation is complete, reattach and secure the electronics cable to J3 on the Acquisition Module, return the electronics enclosure to its original position, replace the bolts, and secure the enclosure in place.
Lifting the meter with the upper enclosure installed but without the bolts installed, may cause the electronics to fall and cause personal injury or equipment damage.
Figure 2-11 Incorrect sling attachment

ever occurs, the hoist ring must be inspected per manufacturer's recommendations prior to any further service. If a proper inspection cannot be performed, discard the hoist ring.
that could damage the ring of the hoist ring.
Use of appropriate safety engineered swivel hoist rings in meter end flanges 37

Section 2: Mechanical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
transducers with the hoist rings. The only exception that safe is to lift the meter with one ASME B16.5 or ASME B16.47 blind flange bolted to each end flange of the meter. NEVER use the hoist rings on the meter to lift other components such as meter tubes, piping or fittings attached to the meter. Doing so will exceed the load rating of the hoist rings.
appropriate case or container per their manufacturer's recommendation.
bolts to keep the tapped holes free of debris and to prevent corrosion.
TA list of approved manufacturers of safety engineered hoist rings is below:
Select an approved supplier from the list below. These vendors can supply the safety- engineered hoist rings. This is not intended to be a complete list.
The appropriate hoist rings can also be purchased directly from Daniel. The following table provides part numbers for reference:
Table 2-1 Hoist ring part number lookup table
Daniel part number1
Hoist ring thread size & load rating 1
American Drill Bushing Co. P/N 1 Carr Lane Manufacturing Co. P/N 1
1-504-90-091 3/8”-16UNC, 1000 lb. 23053 CL-1000-SHR-1
1-504-90-092 1/2”-13UNC, 2500 lb 23301 CL-23301-SHR-1
1-504-90-093 3/4”-10UNC, 5000 lb. 23007 CL-5000-SHR-1
1-504-90-094 1”-8UNC, 10000 lb. 23105 CL-10000-SHR-1
1-504-90-095 1-1/2”-6UNC, 24000 lb. 23202 CL-24000-SHR-1
38 Use of appropriate safety engineered swivel hoist rings in meter end flanges

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 2: Mechanical installation 3-9000-759 Rev D June 2014
To determine the size of the hoist rings required for your meter, use the appropriate table below. Look down the column that matches the ANSI rating of your meter. Find the row that contains your meter size. Follow the row to the end to find the appropriate hoist ring part number.
Table 2-2 Hoist Ring Lookup Table for Daniel 3414 Gas Meters1
ANSI 300 ANSI 600 ANSI 900 ANSI 1500 Daniel Part Number
4” to 10” 4” to 8” 4” to 8” 4” to 6” 1-504-90-091
12” to 18” 10” to 16” 10” to 12” 8” to 10” 1-504-90-092
20” to 24” 18” to 20” 16” to 20” 12” 1-504-90-093
30” to 36” 24” to 30” 24” 16” to 20” 1-504-90-094
36” 30" to 36” 24" to 36” 1-504-90-095
Table 2-3 Hoist Ring Lookup Table for Daniel 3411 or 3412 Gas Meters
ANSI 300 ANSI 600 ANSI 900 ANSI 1500 Daniel Part Number
4” to 10” 4” to 8” 4” to 8” 4” to 6” 1-504-90-091
12” to 18” 10” to 16” 10” to 12” 8” to 10” 1-504-90-092
20” to 24” 18” to 20” 16” to 20” 12” 1-504-90-093
30” to 36” 24” to 30” 24” 16” to 20” 1-504-90-094
36” 30" to 36” 24" to 36” 1-504-90-095
Use of appropriate safety engineered swivel hoist rings in meter end flanges 39

Section 2: Mechanical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
The following instructions are intended to provide general guidelines for using proper lifting slings when lifting a Daniel 3410 Series Gas Ultrasonic Flow Meter by itself. These instructions are intended to be followed in addition to your company's standards or the DOE-STD-1090- 2004 Hoisting and Rigging standard if such company standards do not exist.
slings wrapped around each end of the body as shown below. A choker style sling is recommended.
Figure 2-12 Correct sling attachment

40 Appropriately rated lifting slings

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 2: Mechanical installation 3-9000-759 Rev D June 2014
to the sling manufacturer's procedures for proper inspection of the particular sling you are using.
company's standards for safety factors that must be included when calculating the load rating.
Damage to the meter may occur. If the slings do come in contact with the electronics enclosure, then remove the two bolts holding the enclosure to its base and temporarily remove the head from the meter during the lifting operation (Remove the two bolts holding the enclosure to its base and unplug the cable from the Acquisition Module. Two screws hold this cable in place.) Use a spreader-bar on the slings to prevent contact with the electronics.
on the Acquisition Module, return the electronics enclosure to its original position, replace the bolts, and secure the enclosure in place. Lifting the meter with the upper enclosure installed but with out the bolts installed, may cause the electronics to fall and cause personal injury or electronics damage.
Figure 2-13 Incorrect sling attachment

ever occurs, the slings must be inspected per manufacturer's procedures prior to being placed in any further service.
Appropriately rated lifting slings 41

Section 2: Mechanical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
pipelines
The ambient operating temperature of the Daniel 3410 Series Gas Ultrasonic Flow Meter electronics (i.e. Flameproof enclosure and Intrinsically safe base enclosure) is -40 oF (-40 oC) to +140 oF (+60 oC). If the meter is installed into a pipeline which is heated or cooled outside this temperature range it is necessary to remove the electronics housing from the meter body (i.e. Spool piece acting as process fluid conduit) and mount it next to the meter body on a pipe stand or other rigid structure.
Extended length transducer cables (P/N 2-3-3400-194, 15 ft. long) shall be used to connect the Daniel 3410 Series Gas Ultrasonic Flow Meter electronics to the transducers installed in the meter body. The process temperature must also not exceed the operating temperature range of the transducers. T-11, T-12 and T-21 transducers have an operating range from -4 oF (-20 oC) to 212 oF (+100 oC). T-22 transducers have an operating range from -58 oF (-50 oC) to 212 oF (+100 oC).
The meter body and piping may be extremely hot or cold.
Wear appropriate personal protective equipment when coming in contact with the meter. Failure to do so may result in injury.
42 Mounting requirements in heated or cooled pipelines

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 3: Electrical installation 3-9000-759 Rev D June 2014
3.1 Cable length TTL mode
The maximum cable length is 2000 feet when the Digital Output “TTL” mode is selected.
For the Digital Output “open collector” mode, the maximum cable length depends on the cable parameters, pull-up resistance used, the maximum frequency to output, and frequency input parameters being driven. The following table provides estimated cable lengths for different pull- up resistor values and different Max Frequency settings in the meter using the following cable parameters. The table also provides an estimated cable voltage drop which indicates how much voltage will be across the cabling and effectively indicates to what voltage level the frequency input can be pulled down to by the frequency output.
If the voltage drop is higher than the voltage required for the frequency input to see a low state, then the configuration will most likely not work for your system. Performance of frequency outputs will vary from this table with setup and frequency input being driven.
Table 3-1 Configurations for open collector frequency outputs
Length (2 Conductors) Capacitance Resistance Resistance Frequency Current (2 Conductors)
(x1000ft) nF (Hz) (A) VDC
0.5 16.8 10.00 1000 1016.8 5000 0.024 0.397
1 33.6 20.00 1000 1033.6 1000 0.023 0.780
2 67.2 40.00 1000 1067.2 1000 0.022 1.511
4 134.4 80.00 1000 1134.4 1000 0.021 2.843
0.5 16.8 10.00 500 516.8 5000 0.046 0.780
1 33.6 20.00 500 533.6 5000 0.045 1.511
1.7 57.12 34.00 500 557.12 5000 0.043 2.461
6.5 218.4 130.00 500 718.4 1000 0.033 7.296
The 22 AWG wire characteristics:
Cable length TTL mode 43

Section 3: Electrical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
The meter electronics should be internally grounded for intrinsically safe operations. Connect a wire to the chassis ground lug installed inside the Transmitter Electronics Enclosure as the primary ground. A secondary ground is located outside of the Transmitter Electronics Enclosure (see Figure 3-2).
The internal grounding terminal shall be used as the primary equipment ground. The external terminal is only a supplemental bonding connection where local authorities permit or require such a connection. DO NOT connect digital grounds to the ground lugs.
Figure 3-1 Internal Transmitter Electronics Enclosure chassis ground

A.
A. Transmitter Electronics Enclosure ground lug

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 3: Electrical installation 3-9000-759 Rev D June 2014
Figure 3-2 External ground lug

A.
A. External ground lug
Conduit seals are required for meter installations in hazardous environments. Adhere to safety instructions to protect personnel and equipment.
To reduce the risk of an explosion or fire, conduit runs must have a sealing fitting connected within 457.2 mm (18 inches) of the enclosure. Substitution of components may impair intrinsic safety of the meter.
Failure to keep covers tight during operation may result in serious injury or death.
Substitution of components may impair the intrinsic safety and cause ignition of flammable or combustible atmospheres. Disconnect power before servicing.
Failure to remove power and use Daniel approved components may cause serious injury.
Conduit seals 45

Section 3: Electrical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
Do not open the Transmitter Electronics Enclosure when an explosive gas atmosphere is present. Disconnect equipment from supply circuit before opening the enclosure.
Failure to remove power may result in serious injury or death.
electronics.
shown in see Figure 3-3 and Section 3.5.

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 3: Electrical installation 3-9000-759 Rev D June 2014
Figure 3-3 Electronics field wiring - upper terminal block, switches, ground lug

A.
B.
D.
E.
F.
G.
A. Conduit wiring entry (four entries)
F. Analog In Upper terminal block (continued)
B.
Switches:
Analog In (AO1) Analog Input 1 (Temperature) TT+ TT -
C. Upper terminal block
Analog In (AO2) Analog Input 2 (Pressure)
D. FODO Group 2
PT+ PT -
FODO3 E. Analog Out (Current 4-20mA)
G. Ground lug
Startup for systems using explosion-proof conduit 47

Section 3: Electrical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
Figure 3-4 Transmitter electronics field wiring lower terminal block
A.
B.
D.
E.
F. G. H. A. Upper terminal block
B. FODO Group 1 connections
E. Ethernet
Ethernet (orange and white wire) Ethernet (orange wire) Ethernet (green and white wire) Ethernet (green wire) C. AO1 AO1+ AO1 - D. Serial COMs (RS-232, RS-485) RS-232: RTS, TX, RX, CTS RS-485: TX+, TX -, RX+, RX - (4-wire Full Duplex )
F. 24V loop power (4-24mA inputs/outputs) G. Power In (104VDC - 36VDC) H. Fuse cover

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 3: Electrical installation 3-9000-759 Rev D June 2014
installation information refer to the system wiring diagram (see Appendix A), Daniel MeterLink Software for Gas and Liquid Ultrasonic Meters Quick Start Manual (P/N 3-9000-763) and use the Daniel MeterLink Field Setup Wizard to complete the configuration.
specified by the customer (usually one week) and an electrician has fully tested the connections. After the Acceptance Test is witnessed and approved, seal the conduit.
set in accordance with manufacturer specifications.
Enclosure end caps (see Section 3.7 and Figure 3-11).
(see Section 3.7, Figure 3-12 and Figure 3-13).
Startup for systems using explosion-proof conduit 49

Section 3: Electrical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
Do not open the Transmitter Electronics Enclosure when an explosive gas atmosphere is present. Disconnect equipment from supply circuit before opening the enclosure.
Failure to remove power may result in serious injury or death.
electronics.
installation information refer to the system wiring diagram (see Appendix A), Daniel MeterLink Software for Gas and Liquid Ultrasonic Meters Quick Start Manual (P/N 3-9000-763) and use the Daniel MeterLink Field Setup Wizard to complete the configuration.
specified by the customer (usually one week) and an electrician has fully tested the connections. After the Acceptance Test is witnessed and approved, seal the conduit.
set in accordance with manufacturer specifications.
Enclosure end caps (see Section 3.7 and Figure 3-11).
(see Section 3.7, Figure 3-12 and Figure 3-13).

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 3: Electrical installation 3-9000-759 Rev D June 2014
Daniel MeterLink uses the TCP/IP protocol to communicate with the Daniel 3410 Series Ultrasonic Gas Flow Meter electronics instead of Modbus ASCII or RTU. The TCP/IP protocol only works across either Ethernet, RS-485 full duplex (4-wire) or RS-232. Daniel MeterLink can communicate with multiple meters if they are multi-dropped using 4-wire, full duplex RS-485 mode. The meter electronics is HART capable and provides communication flexibility with Daniel 3410 Series Gas Ultrasonic Flow Meters.
The HART® output provides communication with other field devices (e.g., Field Communicator and AMS™ Device Manager software) and ultimately, communicates key diagnostic information through PlantWeb® architecture.
If not using Ethernet, a full duplex serial connection is necessary for Daniel MeterLink to communicate with a Daniel 3410 Series Gas Ultrasonic Meter.
The Daniel 3410 Series Gas Ultrasonic Flow Meter electronics auto-detects the protocol used and automatically switches between TCP/IP, Modbus ASCII, and Modbus RTU so it is not necessary to make any meter configuration changes to the protocol.
Wiring and I/O 51

Section 3: Electrical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
The meter’s metrology mode and the status of the data transfer from the Acquisition Module to the CPU Module are indicated via light-emitting diode (LED) status indicators. The WRITE PROT. switch protects the meter’s configuration,
Figure 3-5 CPU Module labeling and LED indicators
A. B. C. D. E.
F. G.

A. Acquisition/Measurement mode B. Power C. RX (RS-485/RS-232) - receiving data D. LED 4 - not used E. LED 5 - not used F. TX (RS-485/RS-232) - transmitting data (RS-485 2-wire use TX+ and TX-)
G. Link (Eth1 Link) - user Ethernet connection

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 3: Electrical installation 3-9000-759 Rev D June 2014
CPU Module label or LED Function Switch position indicator or LED
Parity Odd/Even, Stop Bit 1) – RTU (Start Bit 1, Data Bit 8, Parity none, Stop
MEAS System color indicates metrology mode • Acquisition mode
LED 4 • Not used
LED 5 • Not used
RX • RX signal (Port A for RS485 or RS232 ) communication) receiving data
TX • TX signal (Port A for RS485; 2-wire or 4-wire) or RS232 communication) transmitting data
CPU Module labeling and LED indicators 53

Section 3: Electrical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
The Ethernet port IP address, subnet mask, and gateway address are software-configurable. In addition, a meter can be configured to act as a DHCP (Dynamic Host Configuration Protocol) server to assign an IP address to a PC or laptop running Daniel MeterLink. The DHCP server facility is not intended to act as a general purpose DHCP server for a wider network. To this end, no user control is provided over the class or range of IP addresses the unit provides. A standard twisted pair (Cat-5) cable should be used for Ethernet wiring.
It is strongly recommended that the meter be configured using an independent (off-network) single host. After configuration of the Daniel 3410 Series Gas Ultrasonic Flow Meter, the DHCP option must be turned off if used on a LAN/WAN.
Failure to restrict Ethernet and communication access to the Daniel 3410 Series Gas Ultrasonic Flow Meter can result in, among other things, unauthorized access, system corruption, and/or data loss.
User is responsible for ensuring that physical access and Ethernet or electronic access to the Daniel 3410 Series Gas Ultrasonic Flow Meter is appropriately controlled and any necessary security precautions are implemented; such as, establishing a firewall, setting password permissions and/or implementing security levels.

White w/Orange Stripe TX+
Orange w/White Stripe TX -
White w/Green Stripe RX+
Green w/White Stripe RX -
RX+
TX -
TX+
Use ethernet cable, Daniel P/N 3-3400-079, to connect the PC to the meter.
A DIN 41612 48-pin connector is the interface from the CPU Module to the Field Connection Board (male end located on the back of the Field Connection Board).
Ethernet communication · Wire color CPU

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 3: Electrical installation 3-9000-759 Rev D June 2014
If the meter firmware supports Modbus TCP slave functionality, the following controls will be available.
Modbus TCP unit identifier: Enter the Modbus TCP unit identifier here. Valid values are 0-255.
Enable alternate Modbus TCP port: The standard TCP port for Modbus TCP is port 502. This port is always enabled in a meter that supports Modbus TCP. By selecting this option, you can also enable Modbus TCP communications on a secondary TCP port specified by Alternate Modbus TCP port.
Alternate Modbus TCP port:
Enter the alternate TCP port number here after selecting Enable alternate modbus TCP port. Valid port numbers are from 1 to 65535. The meter will not allow some port numbers that are either used by the meter or are defined port numbers for other protocols. Daniel MeterLink™ will prompt you if it was not able to write the specified port number to the meter.
CPU Module labeling and LED indicators 55

Section 3: Electrical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
Use a serial cable, Daniel P/N 3-2500-401, to connect to a PC running Daniel MeterLink. The cable is designed for RS-232 communications which is the serial Port A default configuration (see Appendix A field wiring diagram, Daniel Drawing DMC - 005324). The DB-9 end of the cable plugs directly into the PC running Daniel MeterLink. The three wires on the other end of the cable connect to the CPU Module RS-485/RS-232 terminals. The RED wire goes to RX, the WHITE wire goes to TX, and the BLACK wire goes to COM (see Table 3-6 for Port A wiring). RS-485, 2-wire connection on Port A, uses TX+ and TX- on the CPU Module and has a ground wire.
When Beldon wire No. 9940 or equivalent is used, the maximum cable length for RS-232 communications at 9600 bps is 88.3 meters (250 ft.) and the maximum cable length for RS-485 communication at 57600 bps is 600 meters (1970 ft.).
Port A supports a special override mode which forces the port to use known communication values (19200 baud, address 32, RS-232). Note that the protocol is auto-detected. This mode is expected to be used during meter commissioning (to establish initial communication) and in the event that the user cannot communicate with the meter (possibly due to an inadvertent communication configuration change). Alternately, when using Daniel MeterLink with an Ethernet port, use Ethernet cable, Daniel P/N 3-3400-079, to connect the PC.
Port/Communication Description Common features
Start Bits = 1, Data Bits = 7 2
19200, 38400, 57600, 115000 bps – RTU Protocol:
Start Bits = 1, Data Bits = 82
19200, 38400, 57600, 115000 bps

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 3: Electrical installation 3-9000-759 Rev D June 2014
If not using Ethernet, a full duplex serial connection is necessary for Daniel MeterLink to communicate with a Daniel 3410 Series Gas Ultrasonic Flow Meter.
Figure 3-6 PC to meter serial connection wiring

COM
3.6 I/O connections
The Daniel 3410 Series Gas Ultrasonic Flow Meter provides the I/O connections on the CPU Module.
DB-9 Connector Female · I/O connections 57

Section 3: Electrical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
Figure 3-7 CPU Module I/O connections
A. B. C. D.

A. Frequency/Digital Output 2 B. Frequency/Digital Output 3 C. Analog Output 2 - 4-20mA output D. Analog Input - HART temperature and pressure connections

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 3: Electrical installation 3-9000-759 Rev D June 2014
The meter has three user-configurable selections for configuring either a Frequency output or Digital output (FODO).
Frequency/Digital outputs 59

Section 3: Electrical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 3: Electrical installation 3-9000-759 Rev D June 2014
Frequency/Digital output Source configuration
Frequency /Digital Output 11
FO1A FO1B DO1A DO2B
FODO2
Frequency /Digital Output 22
FODO3
Frequency /Digital Output 32
or · Frequency/Digital outputs 61

Section 3: Electrical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
Output for FODO1 and Digital Output1 (Group 1 on the CPU Module) share a common ground and have 50V isolation. FODO2 and FODO3 (Group 2 on the CPU Module) share a common ground and have 50V isolation. This allows an output to be connected to a different flow computer. The outputs are opto-isolated from the CPU Module and have a withstand voltage of at least 500V rms dielectric.
Figure 3-8 CPU Module - Frequency/Digital outputs common ground

A.
B.
A. FODO1 and Digital input1 - shared common ground (Group 1) B. FODO2 and FODO3 - shared common ground (Group 2)
The Daniel 3410 Series Gas Ultrasonic Flow Meter has the capability to sample analog temperature (Analog Input 1) and pressure (Analog Input 2) with 4-20 mA signals. These analog input signals are configured to sink. The two independent analog input circuits are configured for conventional 4-20 mA service. Also, 24VDC isolated power supply connection is provided for an external power source. Refer to the Field wiring diagram DMC - 005324 in Appendix A.

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 3: Electrical installation 3-9000-759 Rev D June 2014
The Daniel 3410 Series Ultrasonic Gas Flow Meter provides two 4-20 mA analog output signals that are software configurable for either sink or source current (see Appendix A DMC - 005324).
Full HART® functionality is provided so that any commercially available HART® transmitter which meets the specifications of the HART® Communications Foundation can be connected to the Daniel® Liquid Ultrasonic Flow Meter.
The Daniel 3410 Series Gas Ultrasonic Flow Meter provides one digital input that can be used as a general purpose input. The digital input must be configured via the Daniel MeterLink Tools|Edit/Compare Configuration screen.
The meter can be configured to act as a DHCP server. The DHCP server is enabled/disabled via CPU Module DHCP switch as follows:
CPU Module switch DHCP server disabled

Analog output settings 63
DHCP server enabled · DHCP Switch 2

Section 3: Electrical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
The meter’s configuration parameters and firmware can be protected against changes via the CPU Module WRITE PROT. switch as follows:
CPU Module switch Configuration protected
WRITE PROT.
Configuration unprotected · ON (default setting) · Switch 3

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 3: Electrical installation 3-9000-759 Rev D June 2014
Located inside the Transmitter Electronics Enclosure is a connector for a user-provided external power source, a 2 Amp fuse and a 24V loop power connection for ultrasonic meter analog outputs, temperature transmitter or pressure transmitter devices. The current is limited to 88mA.
Figure 3-9 CPU Module power source connections
A. B.

A. Power In connector (main power) B. 24V LOOP POWER C. 2 Amp fuse (used for the main power input)
External power source connection and fuse 65

Section 3: Electrical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
Security seals protect the integrity of the meter metrology and prevent tampering with transducer assemblies. The following sections detail how to properly seal the Daniel 3410 Series Gas Ultrasonic Flow Meter after commissioning. The security seal wires are commercially available.
Be sure to set the WRITE PROT. switch on the CPU Module to the ON position prior to sealing the enclosure.
Use the following instructions to install the security seal wires on the Transmitter Electronics Enclosure.
Figure 3-10 Transmitter electronics enclosure security latch
A.
B.
A. Transmitter Electronics Enclosure end cap. Optional glass endcap for Local Display B. Security latch

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 3: Electrical installation 3-9000-759 Rev D June 2014
Choose holes that minimize counterclockwise rotation of the end cap when the security wire is taut (maximum wire diameter .078 inch; 2.0mm).
Figure 3-11 Transmitter Electronics Enclosure security seals

A.
B.
A. Transmitter Electronics Enclosure end cap B. Security wire seals
Seal Transmitter Electronics Enclosure 67

Section 3: Electrical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
Use the following instructions to install the security seal wire on the Base Enclosure.
Figure 3-12 Base Enclosure wire seal installation

A.
B.
A. Base Enclosure cover
B. Security wire seals
is taut.

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 3: Electrical installation 3-9000-759 Rev D June 2014
adjacent socket head screw. Twist the wire, removing all slack and seal.
Figure 3-13 Base Enclosure security seals

A.
B.
D.
A. Transmitter Electronics Enclosure
B. Security wire seals e C. Transmitter Electronics endcap security latch
D. Base Enclosure
Base Enclosure Security Seals 69

Section 3: Electrical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
Use the following instructions and Figure 3-14 to install the security seal wire on the transducer assembly.
(Item B) and feed through one of the two holes in the transducer cable nut (Item A). Note: Choose holes that minimize counterclockwise rotation of the transducer cable nut when the security wire (Item C) is taut.
Figure 3-14 Transducer assembly security seal

A.
B.
A. Transducer cable nut B. Transducer cable connector C. Security wire seal

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 3: Electrical installation 3-9000-759 Rev D June 2014
The unit should be properly sealed with a sealing compound after electrical connections have been tested according to the customer's Best Practices schedule. Some areas require a witnessed Acceptance Test for the installed system and require that the meter run for a predetermined length of time (approximately one to two weeks) before the unit is sealed. This allows time to verify all electrical connections are correct, that the meter is accurately measuring flow and that the meter meets the customer’s installation requirements. See Section 3.4.1 and Section 3.4.2
Sealing the unit 71

Section 3: Electrical installation Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 4: Configuration 3-9000-759 Rev D June 2014
After the mechanical and electrical installation is complete and connectivity is established, use the Daniel MeterLink Software for Gas and Liquid Ultrasonic Meters Quick Start Manual (P/N 3-9000-763) to setup software communications with the meter.
Vista®, Windows® 7 or Windows® 8).
MeterLink communications.
Registration Wizard to keep you informed about updates and technical support. Registration options include:
Company name, display units, Liquid Meter volume units and other interface settings)
and setup the connections properties.
document the meter’s “As Found” settings.
Daniel MeterLink Setup 73

Section 4: Configuration Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
units) volume units, flow rate time, low flow cutoff, contract hour, enable reverse flow alarm, set meter time and notepad comments. Select Next to continue to Frequency/ Digital Outputs page.
status. Select the Source for each Frequency/Digital output and select the desired drive Mode. The Mode options are Open Collector which requires an external excitation voltage and pull-up resistor or TTL mode which outputs a 0-5 VDC signal. Select Next to continue to Frequency Outputs page.
B phase, maximum frequency output (Hertz) and full scale volumetric flow rate. Select Next to continue to Meter Digital Outputs.
Digital output 2A and Digital output 2B based on Frequency validity or flow direction.
if the output of the ultrasonic meter is reversed from what a flow computer is expecting, select Inverted Operation. This changes the digital output from a HIGH for a TRUE condition to output a LOW for a TRUE condition. Select Next to continue to Current Outputs.
(Forward, Reverse or Absolute) and Full scale volumetric flow rate used with output (20mA maximum). Alarm action parameters determines the state the output will drive during an alarm condition (High 20mA, Low - 4 mA, Hold last value, Very low - 3.5, Very high 20.5 mA or None). Select Next to configure the HART® Output(s) parameters.
Secondary, Third and Fourth variable. The Primary variable is set to match the Content set for Current output 1. If a second current output is available, the Secondary variable is set to match the Content set for Current output 1) Identification and HART® units (volume units, flow rate time units, velocity units, pressure and temperature units). Select Next to continue to the Meter Corrections page.
74 Field Setup Wizard

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 4: Configuration 3-9000-759 Rev D June 2014
pressure and linear expansion corrections, pipe outside diameter The Young’s Modulus value (ratio of tensile stress to tensile strain) and the Poisson’s ratio value (the absolute ratio of the pipe material lateral strain over axial strain) for 3412 and 3411 Gas Ultrasonic Flow meters. Click Next to continue to the Temperature and Pressure page.
alarm limits for both. The alarm limit selections are hold last output value or use fixed value.
Click Next to continue to the Gas Chromatograph Setup page.
chromatograph.
Field Setup Wizard 75

Section 4: Configuration Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
displayed for Daniel Gas Ultrasonic meters if both temperature and pressure are set to Live or Fixed and Base condition correction is selected on the Startup Page. Configuration parameters include:
Click Next to continue to the Continuous Flow Analysis page, if View Continuous Flow Analysis setup was selected on the Startup page.
Click Next to continue to the Local display page, if View local display setup was selected on the Startup page.
Items list box and select or modify the parameters that will be displayed; the Display items, the Display units and the Scroll delay.
76 Field Setup Wizard

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 4: Configuration 3-9000-759 Rev D June 2014
Display Items:
The Local Display’s labels and descriptions are shown below:
Local display labels, descriptions and valid units
QFLOW — Uncorrected volume flow rate
TDYVL — Current day’s forward uncorrected volume
TDYVL — Current day’s reverse uncorrected volume
YSTVL — Previous day’s forward uncorrected volume
YSTVL — Previous day’s reverse uncorrected volume
TOTVL — Forward uncorrected volume
TOTVL — Reverse uncorrected volume
QBASE — Corrected volume flow rate
Field Setup Wizard 77

Section 4: Configuration Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
Local display labels, descriptions and valid units
TDYVL — Current days forward corrected volume
TDYVL — Current days reverse corrected volume
YSTVL — Previous days forward corrected volume
YSTVL — Previous days reverse corrected volume
TOTVL — Forward corrected volume
TOTVL — Reverse corrected volume
VEL — Average flow velocity
SOS — Average sound velocity
TEMP — Flow-condition temperature
PRESS — Flow-condition pressure
FRQ1A — Frequency channel 1A
FRQ1B — Frequency channel 1B
78 Field Setup Wizard

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 4: Configuration 3-9000-759 Rev D June 2014
Local display labels, descriptions and valid units
KFCT1 — Frequency 1 K-factor
FRQ2A — Frequency channel 2A
FRQ2B — Frequency channel 2B
KFCT2 — Frequency 2 K-factor
AO1 — Analog Output 1 current
AO2 — Analog Output 2 current
Display units:
The Meter volume units displayed are either U.S. Customary or Metric. To modify the Display Units, configure the Meter units system in the Field Setup Wizard — General Page.
– Thousand cubic feet
Field Setup Wizard 79

Section 4: Configuration Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
Scroll delay:
The Scroll Delay is the time interval for the selected display items to be shown on the Local Display. The default scroll delay setting is five seconds. Click the spin box up or down arrow to increase or decrease the length of time an item displays.
document the “As Left” settings.
meter
This procedure assumes you have AMS Device Manager installed on the host computer and have downloaded the latest Daniel Liquid Ultrasonic Meter Device Description (DD).
If not installed, click the link below to download the AMS device installation tool kit.
http://www2.emersonprocess.com/en-US/documentation/deviceinstallkits/Pages/devicein- stallkitsearch.aspx
80 Using AMS Device Manager to configure the meter

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 4: Configuration 3-9000-759 Rev D June 2014
Procedure - installing AMS Device Description
Figure 4-1 AMS Device Description search

Using AMS Device Manager to configure the meter 81

Section 4: Configuration Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
displays. Click the Save button to save the files to your host system. You may use the default download location or change the directory.

Figure 4-2 AMS file download complete

82 Using AMS Device Manager to configure the meter

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 4: Configuration 3-9000-759 Rev D June 2014
Figure 4-3 AMS Device Manager

Figure 4-4 AMS Device Manager - Overview

Using AMS Device Manager to configure the meter 83

Section 4: Configuration Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
AMS Device Manager - Guided Setup
The Guided setup wizard provides configuration parameter settings for the meter. The Guided Setup is a subset of the Manual Setup parameters.
Figure 4-5 AMS Device Manager - Guided Setup

Volume units, Flow rate time units, Velocity units, Pressure units and Temperature units. Click Apply to write the parameters to the meter.
Frequency/Digital outputs, Frequency and Digital Outputs 1 and 2, Analog outputs, Digital Input, Pressure and Temperature.
a. Analog output 1 (HART) - Content (Primary Variable) displays Uncorrected Flow Rate and is a read only attribute). Configure Direction (flow), Lower Range value, Upper range value and Alarm Action and view the HART Parameters Tag, Date, Descriptor, Message, Final Assembly Number Poll Address, Number of Response Preambles. b. Analog Output 2 - Content (Secondary Variable) displays Uncorrected Flow Rate and has a read only attribute. Configure Direction (flow), Lower Range value, Upper range value and Alarm Action. Map the Third and Four variables using the Manual Setup wizard. Selections include Uncorrected Volume Flow Rate, Pressure and Temperature.
84 Using AMS Device Manager to configure the meter
Procedure

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 4: Configuration 3-9000-759 Rev D June 2014
meter.
a. Click the Frequency/Digital Outputs tab to configure Frequency/Digital Output 1, 2 and 3 Source and drive Mode. Select the Source for each Frequency/Digital output and select the desired drive Mode. The Mode options are Open Collector which requires an external excitation voltage and pull-up resistor or TTL mode which outputs a 0-5 VDC signal (each Frequency output has an A and B output phase). (Refresh Note: If changes are made to any Source variable on this page, apply the changes and navigate to the Guided Setup page. Navigate back to the Manual Setup for the changes to be reflected in other Manual Setup pages). b. Click the Frequency and Digital Output 1 tab to configure the Content, (flow) Direction, Channel B Phase frequency output, Lag forward, Lead Reverse or Lead Forward, Lag Reverse (Phase B lags Phase A while reporting forward flow and lead Phase A while reporting reverse flow or the opposite), Digital Output 1 Channel A Content and Polarity, Channel B Content and Polarity, Maximum Frequency, and Lower and Upper Range Units of Measure. c. Click the Frequency and Digital Output 2 tab and repeat Step 3b to configure Frequency and Digital Output 2 parameters.
text, Final Assembly number, Poll address and number of response preambles are displayed). After all of the data is entered click Apply to write the parameters to the meter.
Reverse Flow. Click the OK button to return to the Overview page.
Data, Path Information, Flow Totals, and All Variables data is populated after you are connected to the meter.
a. Click the Flow Data tab and view the Flow Direction (Forward or Reverse), Average Flow and Average Sound Velocities values. b. Click the Path Information tab and view the Chord performance, Gain, SNR (Signal to Noise Ratio) Signal strength (mV), and Noise (mV). c. Click the Flow Totals tab to view the volume totals (forward and reverse uncorrected volume).
Using AMS Device Manager to configure the meter 85

Section 4: Configuration Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
d. Click the All Variables tab to view a graphical display of the Primary, Secondary, Third and Fourth Variables.
Figure 4-6 AMS Device Manager - Service Tools All Variables status indicators

configuration.
values calculated from the Full scale volumetric flow rate used with frequency outputs and the Maximum frequency for frequency output.
Figure 4-7 Display Meter K-Factors

Click Next to return to the Device Manager Overview page.
86 Using AMS Device Manager to configure the meter

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 4: Configuration 3-9000-759 Rev D June 2014
AMS Device Manager - Manual Setup
Use the Manual Setup wizard to configure the meter’s parameters. See Figure 4-3 and Figure 4-4 and from the AMS Device Manager Configure menu click Manual Setup.
Figure 4-8 AMS Device Manager - Configure Manual Setup

only and are configured for Uncorrected Flow Rate. The Third and Fourth variable configuration choices include Pressure and Temperature.
Manager - Guided Setup, Step 3. The read only Secondary variable Content, Uncorrected Flow Rate, displays. Use the drop-down arrow and select the (flow) Direction - Forward or Reverse. Enter a Lower and Upper Range limit. Set the Alarm Action parameters. Click Apply, after you enter the data to write the parameters to the meter.
Using AMS Device Manager to configure the meter 87
Procedure

Section 4: Configuration Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
AMS Device Manager - Guided Setup, Step 4a.). (Refresh Note: If changes are made to any Source variable on this page, apply the changes and navigate to the Guided Setup page. Navigate back to the Manual Setup for the changes to be reflected in other Manual Setup pages). Click Apply, after you enter the data to write the parameters to the meter.
AMS Device Manager - Guided Setup, Step 4b. Click Apply, after you enter the data to write the parameters to the meter.
Manager - Guided Setup, Step 4c to configure the Frequency and Digital Output 2 parameters. Click Apply, after you enter the data to write the parameters to the meter.
Analog or Fixed), Min and Max input limits corresponding to 4 mA and 20 mA respectively and the Low and High alarm limits. Click Apply, after you enter the data to write the parameters to the meter.
Fixed), Min and Max input limits corresponding to 4 mA and 20 mA respectively and the Low and High alarm limits. Select either Gage or Absolute for the type of pressure reading desired. If a live pressure transmitter is connected, select the type of reading the transmitter outputs. If Absolute is selected, you must also enter the Atmospheric pressure. Click Apply, after you enter the data to write the parameters to the meter.
purpose or set to Inverted when used for calibration. Click Apply, after you choose the calibration data to write the parameters to the meter.
a. Calibration Polarity configuration parameter selections are: – Digital Input 1 Calibrate Active High – Digital Input 1 Calibrate Active Low
88 Using AMS Device Manager to configure the meter

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 4: Configuration 3-9000-759 Rev D June 2014
b. Calibration Gating configuration parameter selections are: – Edge gated, active high
Figure 4-9 Gating configuration parameter Edge gated, active high

Calibration Start Calibration Stop
– Edge gated, active low
Figure 4-10 Gating configuration parameter Edge gated, active low
Calibration Start Calibration Stop
– State gated, active high
Figure 4-11 Gating configuration parameter State gated, active high

Calibration Start Calibration Stop
– State gated, active low
Figure 4-12 Gating configuration parameter State gated, active low

Calibration Start Calibration Stop
Using AMS Device Manager to configure the meter 89

Section 4: Configuration Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
Figure 4-13 Configure Flow Analysis Alert

default setting is Disabled. Click the Disabled button to send the feature command to the meter. Check for a response error. If no error response is received, click the Enable button.
a. Enter the minimum reverse flow velocity above which to accumulate flow in the reverse direction for this alert. Enter a positive value for the Reverse Flow Zero Cutoff. Click the Next button to write the values to the meter. Check for an error response. If no error response is received, click the Next button. The Detect Reverse Flow enabled page displays. Click the Next button to display Detect Reverse Flow disabled. b. If an error message is returned, click the Next button to display the Method Complete page. c. Click the Set Flow Range Limits button and enter a positive value for the Flow Analysis Lower Velocity Range and the Upper Velocity Range Limits. When the velocity is outside of the limit parameters, an alert is triggered. Click the Next button to display the Method Complete page.
90 Using AMS Device Manager to configure the meter

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 4: Configuration 3-9000-759 Rev D June 2014
statuses or to edit the configuration parameters.
a. Click the Service Tools|Alerts tab. If an alert condition exists, the alert type and description displays. Recommended actions are listed to assist you in a resolution. After you resolve the alert condition, click the Acknowledge button to clear the alert. Click Apply to write the changes to the meter. If no alert condition is active, click OK to close the device window.
Figure 4-14 AMS Device Manager - Service Tools Alerts

b. If you change the device configuration, a confirmation dialog displays and prompts you to write the changes to the meter. Click Yes to write the changes to the meter or click No to cancel pending changes.
Figure 4-15 Configuration changes dialog

Using AMS Device Manager to configure the meter 91

Section 4: Configuration Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
c. Click the Service Tools|Variables tab. The Variables page displays tabs for the device’s Flow Data, Path Information, Flow Totals, and All Variables).
Figure 4-16 AMS Device Manager - Service Tools

d. The Service Tools|Flow Data page includes charts for flow and sound velocities. The flow values (flow direction, average flow velocity and average sound velocity) parameters are displayed for the connected device. e. Click Service Tools|Variables|Path Information tab to view the device’s chord performance (%), Gain (dB), SNR (dB), Signal (mV) and Noise (mV). f. Click Service Tools|Variables|Flow Totals to view the volume totals (Forward and reverse Uncorrected Volume) parameters for the connected device.
92 Using AMS Device Manager to configure the meter

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 4: Configuration 3-9000-759 Rev D June 2014
g. Click Service Tools|Variables|All Variables tab to view Primary, Secondary, Third and Fourth Variable parameter status.
Figure 4-17 AMS Device Manager - Service Tools All Variables
Gauges display each variable’s status as good or bad. If a status is bad refer to the Service Tools Alerts page for recommended actions to resolve the alert condition. Also refer to the Field Device Specification manual (P/N 3-9000-762) for Commands 48 and 140 details.
Alerts are triggered for Command 48 Additional device status and Command 140 detailed status information. Alerts are grouped as Failed - Fix Now, Maintenance - Fix Soon and Advisory according to the severity level; 1-6. Severity 1 is the highest and 6 is the lowest level.
Using AMS Device Manager to configure the meter 93

Section 4: Configuration Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
h. Click the Service Tools|Trends tab to display the device variables (uncorrected volume flow rate, pressure and temperature) trends.
Figure 4-18 AMS Device Manager - Service Tools Trends

Primary and Secondary variables display real-time uncorrected volume flow rate
trends. The third and fourth variables charts displays trends for temperature and pressure.
digital to analog trim adjustment of the first milliampere output. The 4mA and 20mA output current values should equal the plant’s standard values. Click Yes to confirm the configuration changes. Repeat this step to trim Analog Output 2 current. Click Apply to write the output trim values to the meter. Click OK to navigate back to the Service Tools page.
following:
a. Enable the Write Protect switch on the CPU Module to protect the meter’s configuration. b. Replace the end cap and if required, apply security seals through the endcap holes and through the hex head bolts that secure the Bracket/Cover to the Base enclosure.
94 Using AMS Device Manager to configure the meter

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 4: Configuration 3-9000-759 Rev D June 2014
meter
Follow all guidelines and precautions described in the Field Communicator User Manual and in the 3818 LNG Liquid Ultrasonic Flow Meter documentation when working in a hazardous area.
http://www2.emersonprocess.com/en-US/brands/Field-Communicator/Pages/Documenta- tion.aspx
http://www2.emersonprocess.com/en-US/brands/daniel/Flow/ultrasonics/Pages/Ultra- sonic.aspx
Using a Field Communicator to configure the meter 95
Installation Requirements

Section 4: Configuration Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
instructions provided with your handheld device. Register the product to activate the end user license.
the battery in a hazardous area environment. The power supply is not intrinsically safe
electronics enclosure.
Figure 4-19 3414 transmitter field wiring conduit entries
A.
A. Field wiring conduit entries (4)
96 Using a Field Communicator to configure the meter
Procedure

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 4: Configuration 3-9000-759 Rev D June 2014
Appendix A, drawing DMC-005324.
Figure 4-20 Field Communicator wiring diagram for the 3410 Series electronics


navigate through the device menus.
manual (P/N 3-9000-762) for the device fast key sequences. Included in the menu tree are:
manual or the contacts included in the Field Communicator User’s Manual.
Using a Field Communicator to configure the meter 97

Section 4: Configuration Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
For the integrity of the meter metrology and to prevent tampering with the transmitter electronics and transducer assemblies, attach security latches on the end caps and install security wires on the Transmitter Electronics Enclosure end caps, the Bracket/Cover cap head screws. See Section 3.7 and Section 3.7.4.
Seal the conduit ports with sealing compound according to the customer’s requirements (e.g., after approximately one to two weeks of run time). Also, see Section 3.4.1.
98 Security seals for the meter (optional)

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Engineering drawings 3-9000-759 Rev D June 2014
A.1 3410 Series engineering drawings
List of Engineering Drawings
This appendix contains the following engineering drawing(s) for the ultrasonic meter:
DMC-005324 Daniel 3410 Series Ultrasonic Gas Flow Meter System Wiring Diagram
3410 Series engineering drawings 99

Engineering drawings Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
100 3410 Series engineering drawings









Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Open source licenses 3-9000-759 Rev D June 2014
Source code for executable files or libraries included in this product is provided per the indicated license in the table below. Hyperlinks to the controlling organization's websites are included in Section B.1 through Section B.4.
Table B-1 Open source licences
Package File specification License Summary
base_libs-1.2-1 base_libs LGPL Base Libraries (from toolchain)
busybox-1.1.3-1 busybox GPL A small executable that replaces many UNIX utilities
dev-1.1-1 dev GPL Device files for a small embedded system
devmem2-1.0-1 devmem2 GPL Simple program to read/write from/to any location
ethtool-3-1 ethtool GPL Ethernet settings tool for PCI Ethernet cards
expat-2.0.1-1 expat MIT XML 1.0 parser
fake-provides-1.0-5 fake-provides GPL Fake provides to satisfy package dependencies
gdb-6.6cs-1 gdb GPL Gdb - GNU Source level debugger for C, C++
kernel-2.6.37-6 kernel-2.6.37- mpc8313erd
GPL Linux kernel (core of the Linux operating system)
libpcap-0.8.3-1 libpcap BSD A system-independent interface for user-level pa
libtermcap-2.0.8-31_1 libtermcap LGPL A basic system library for accessing the termcap
lwIP BSD A lightweight TCP/IP stack
merge-0.1-1 merge GPL Merge files for an embedded root filesystem
modeps-1.0-1 modeps GPL Generate module dependency file
mtd-utils-20060302-1 mtd-utils GPL Memory Technology Device tools
net-tools-1.60-1 net-tools GPL Basic networking tools
ppp-2.4.4-1 ppp BSD Like a Point-to-Point Protocol daemon
skell-1.16-2 skell GPL Skelleton files for an embedded root filesystem
sqlite-3.6.22-1 sqlite Public domain SQLite is a C library that imple- ments an embeddable SQL database

Open source licenses Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
Package File specification License Summary
strace-4.5.14-1 strace BSD trace system calls associated with a running pro
sysconfig-1.2-1 sysconfig GPL System configuration package
sysfsutils-2.1.0-1 sysfsutils GPL/LGPL sysfs utilities
tcpdump-3.8.3-1 tcpdump BSD A network traffic monitoring tool
termcap-1.2-1 termcap BSD minimal /etc/termcap needed by minicom etc
u-boot-1.3.0-1 u-boot-1.3.0- mpc8313erdb
GPL Universal Bootloader firmware
ubi-utils-1.4.2-1 ubi-utils GPL Tools for maintaining Unsorted Block Image Device
vsftpd-2.2.2-1 vsftpd GPL vsftpd - Very Secure Ftp Daemon
zlib-1.2.3-2 zlib zlib Distribution zlib compression utilities and libraries
Follow the link below to the Daniel® Liquid Ultrasonic Products GPL webpage for additional open source information and zipped source code files.
For more details about GNU GPL (General Public License), follow the link below:
Daniel® Measurement and Control, Inc., uses GPL version 2.
http://www.gnu.org/licenses/old-licenses/gpl-2.0.html
The GNU GPL is currently version 3
http://www.gnu.org/licenses/quick-guide-gplv3.html
For older versions of the GNU General Public License, follow the link below:
http://www.gnu.org/licenses/old-licenses/old-licenses.html#GPL
See GPL license on the following pages.
102 GNU General Public License

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Open source licenses 3-9000-759 Rev D June 2014
Copyright (C) 1989, 1991 Free Software Foundation, Inc.
59 Temple Place, Suite 330, Boston, MA 02111-1307 USA
Everyone is permitted to copy and distribute verbatim copies
of this license document, but changing it is not allowed.
The licenses for most software are designed to take away your freedom to share and change it. By
contrast, the GNU General Public License is intended to guarantee your freedom to share and change
free software--to make sure the software is free for all its users. This General Public License applies to
most of the Free Software Foundation's software and to any other program whose authors commit to
using it. (Some other Free Software Foundation software is covered by the GNU Library General Public
License instead.) You can apply it to your programs, too.
When we speak of free software, we are referring to freedom, not price. Our General Public Licenses
are designed to make sure that you have the freedom to distribute copies of free software (and
charge for this service if you wish), that you receive source code or can get it if you want it, that you
can change the software or use pieces of it in new free programs; and that you know you can do
these things.
To protect your rights, we need to make restrictions that forbid anyone to deny you these rights or to
ask you to surrender the rights. These restrictions translate to certain responsibilities for you if you
distribute copies of the software, or if you modify it.
GNU General Public License 103
Preamble

Open source licenses Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
For example, if you distribute copies of such a program, whether gratis or for a fee, you must give the
recipients all the rights that you have. You must make sure that they, too, receive or can get the
source code. And you must show them these terms so they know their rights.
We protect your rights with two steps: (1) copyright the software, and (2) offer you this license which
gives you legal permission to copy, distribute and/or modify the software.
Also, for each author's protection and ours, we want to make certain that everyone understands that
there is no warranty for this free software. If the software is modified by someone else and passed
on, we want its recipients to know that what they have is not the original, so that any problems
introduced by others will not reflect on the original authors' reputations.
Finally, any free program is threatened constantly by software patents. We wish to avoid the danger
that redistributors of a free program will individually obtain patent licenses, in effect making the
program proprietary. To prevent this, we have made it clear that any patent must be licensed for
everyone's free use or not licensed at all.
The precise terms and conditions for copying, distribution and modification follow.
104 GNU General Public License

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Open source licenses 3-9000-759 Rev D June 2014
holder saying it may be distributed under the terms of this General Public License. The "Program",
below, refers to any such program or work, and a "work based on the Program" means either the
Program or any derivative work under copyright law: that is to say, a work containing the Program or
a portion of it, either verbatim or with modifications and/or translated into another language.
(Hereinafter, translation is included without limitation in the term "modification".) Each licensee is
addressed as "you".
Activities other than copying, distribution and modification are not covered by this License; they are
outside its scope. The act of running the Program is not restricted, and the output from the Program
is covered only if its contents constitute a work based on the Program (independent of having been
made by running the Program). Whether that is true depends on what the Program does.
medium, provided that you conspicuously and appropriately publish on each copy an appropriate
copyright notice and disclaimer of warranty; keep intact all the notices that refer to this License and
to the absence of any warranty; and give any other recipients of the Program a copy of this License
along with the Program.
You may charge a fee for the physical act of transferring a copy, and you may at your option offer
warranty protection in exchange for a fee.
GNU General Public License 105

Open source licenses Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
on the Program, and copy and distribute such modifications or work under the terms of Section 1
above, provided that you also meet all of these conditions:
a) You must cause the modified files to carry prominent notices stating that you changed the
files and the date of any change.
b) You must cause any work that you distribute or publish, that in whole or in part contains or
is derived from the Program or any part thereof, to be licensed as a whole at no charge to
all third parties under the terms of this License.
c) If the modified program normally reads commands interactively when run, you must cause
it, when started running for such interactive use in the most ordinary way, to print or
display an announcement including an appropriate copyright notice and a notice that there
is no warranty (or else, saying that you provide a warranty) and that users may
redistribute the program under these conditions, and telling the user how to view a copy
such an announcement, your work based on the Program is not required to print an
announcement.)
These requirements apply to the modified work as a whole. If identifiable sections of that work are not
derived from the Program, and can be reasonably considered independent and separate works in
themselves, then this License, and its terms, do not apply to those sections when you distribute them
as separate works. But when you distribute the same sections as part of a whole which is a work
based on the Program, the distribution of the whole must be on the terms of this License, whose
permissions for other licensees extend to the entire whole, and thus to each and every part regardless
of who wrote it.
Thus, it is not the intent of this section to claim rights or contest your rights to work written entirely
by you; rather, the intent is to exercise the right to control the distribution of derivative or collective
works based on the Program.
106 GNU General Public License

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Open source licenses 3-9000-759 Rev D June 2014
In addition, mere aggregation of another work not based on the Program with the Program (or with a
work based on the Program) on a volume of a storage or distribution medium does not bring the other
work under the scope of this License.
executable form under the terms of Sections 1 and 2 above provided that you also do one of the
following:
a) Accompany it with the complete corresponding machine-readable source code, which must
be distributed under the terms of Sections 1 and 2 above on a medium customarily used
for software interchange; or,
b) Accompany it with a written offer, valid for at least three years, to give any third party, for
a charge no more than your cost of physically performing source distribution, a complete
machine-readable copy of the corresponding source code, to be distributed under the
terms of Sections 1 and 2 above on a medium customarily used for software interchange;
c) Accompany it with the information you received as to the offer to distribute corresponding
source code. (This alternative is allowed only for noncommercial distribution and only if
you received the program in object code or executable form with such an offer, in accord
with Subsection b above.)
The source code for a work means the preferred form of the work for making modifications to it. For
an executable work, complete source code means all the source code for all modules it contains, plus
any associated interface definition files, plus the scripts used to control compilation and installation of
the executable. However, as a special exception, the source code distributed need not include
anything that is normally distributed (in either source or binary form) with the major components
(compiler, kernel, and so on) of the operating system on which the executable runs, unless that
component itself accompanies the executable.
GNU General Public License 107
or,

Open source licenses Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
If distribution of executable or object code is made by offering access to copy from a designated
place, then offering equivalent access to copy the source code from the same place counts as
distribution of the source code, even though third parties are not compelled to copy the source along
with the object code.
under this License. Any attempt otherwise to copy, modify, sublicense or distribute the Program is
void, and will automatically terminate your rights under this License. However, parties who have
received copies, or rights, from you under this License will not have their licenses terminated so long
as such parties remain in full compliance.
grants you permission to modify or distribute the Program or its derivative works. These actions are
prohibited by law if you do not accept this License. Therefore, by modifying or distributing the
Program (or any work based on the Program), you indicate your acceptance of this License to do so,
and all its terms and conditions for copying, distributing or modifying the Program or works based on
it.
automatically receives a license from the original licensor to copy, distribute or modify the Program
subject to these terms and conditions. You may not impose any further restrictions on the recipients'
exercise of the rights granted herein. You are not responsible for enforcing compliance by third
parties to this License.
reason (not limited to patent issues), conditions are imposed on you (whether by court order,
agreement or otherwise) that contradict the conditions of this License, they do not excuse you from
the conditions of this License. If you cannot distribute so as to satisfy simultaneously your obligations
under this License and any other pertinent obligations, then as a consequence you may not distribute
the Program at all. For example, if a patent license would not permit royalty-free redistribution of the
108 GNU General Public License

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Open source licenses 3-9000-759 Rev D June 2014
Program by all those who receive copies directly or indirectly through you, then the only way you
could satisfy both it and this License would be to refrain entirely from distribution of the Program.
If any portion of this section is held invalid or unenforceable under any particular circumstance, the
balance of the section is intended to apply and the section as a whole is intended to apply in other
circumstances.
It is not the purpose of this section to induce you to infringe any patents or other property right
claims or to contest validity of any such claims; this section has the sole purpose of protecting the
integrity of the free software distribution system, which is implemented by public license practices.
Many people have made generous contributions to the wide range of software distributed through
that system in reliance on consistent application of that system; it is up to the author/donor to decide
if he or she is willing to distribute software through any other system and a licensee cannot impose
that choice.
This section is intended to make thoroughly clear what is believed to be a consequence of the rest of
this License.
by copyrighted interfaces, the original copyright holder who places the Program under this License
may add an explicit geographical distribution limitation excluding those countries, so that distribution
is permitted only in or among countries not thus excluded. In such case, this License incorporates the
limitation as if written in the body of this License.
License from time to time. Such new versions will be similar in spirit to the present version, but may
differ in detail to address new problems or concerns.
Each version is given a distinguishing version number. If the Program specifies a version number of
this License which applies to it and "any later version", you have the option of following the terms and
GNU General Public License 109

Open source licenses Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
conditions either of that version or of any later version published by the Free Software Foundation. If
the Program does not specify a version number of this License, you may choose any version ever
published by the Free Software Foundation.
conditions are different, write to the author to ask for permission. For software which is copyrighted
by the Free Software Foundation, write to the Free Software Foundation; we sometimes make
exceptions for this. Our decision will be guided by the two goals of preserving the free status of all
derivatives of our free software and of promoting the sharing and reuse of software generally.
PROGRAM, TO THE EXTENT PERMITTED BY APPLICABLE LAW. EXCEPT WHEN OTHERWISE STATED IN
WRITING THE COPYRIGHT HOLDERS AND/OR OTHER PARTIES PROVIDE THE PROGRAM "AS IS"
WITHOUT WARRANTY OF ANY KIND, EITHER EXPRESSED OR IMPLIED, INCLUDING, BUT NOT
LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR
PURPOSE. THE ENTIRE RISK AS TO THE QUALITY AND PERFORMANCE OF THE PROGRAM IS WITH
YOU. SHOULD THE PROGRAM PROVE DEFECTIVE, YOU ASSUME THE COST OF ALL NECESSARY
SERVICING, REPAIR OR CORRECTION.
COPYRIGHT HOLDER, OR ANY OTHER PARTY WHO MAY MODIFY AND/OR REDISTRIBUTE THE
PROGRAM AS PERMITTED ABOVE, BE LIABLE TO YOU FOR DAMAGES, INCLUDING ANY GENERAL,
SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES ARISING OUT OF THE USE OR INABILITY TO
USE THE PROGRAM (INCLUDING BUT NOT LIMITED TO LOSS OF DATA OR DATA BEING RENDERED
INACCURATE OR LOSSES SUSTAINED BY YOU OR THIRD PARTIES OR A FAILURE OF THE PROGRAM
TO OPERATE WITH ANY OTHER PROGRAMS), EVEN IF SUCH HOLDER OR OTHER PARTY HAS BEEN
ADVISED OF THE POSSIBILITY OF SUCH DAMAGES.
110 GNU General Public License

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Open source licenses 3-9000-759 Rev D June 2014
How to Apply These Terms to Your New Programs
If you develop a new program, and you want it to be of the greatest possible use to the public, the
best way to achieve this is to make it free software which everyone can redistribute and change under
these terms.
To do so, attach the following notices to the program. It is safest to attach them to the start of each
source file to most effectively convey the exclusion of warranty; and each file should have at least the
"copyright" line and a pointer to where the full notice is found.
One line to give the program's name and a brief idea of what it does.
Copyright (C)
This program is free software; you can redistribute it and/or modify it under the terms of the
GNU General Public License as published by the Free Software Foundation; either version 2 of
the License, or (at your option) any later version.
This program is distributed in the hope that it will be useful, but WITHOUT ANY WARRANTY;
without even the implied warranty of MERCHANTABILITY or FITNESS FOR A PARTICULAR
PURPOSE. See the GNU General Public License for more details.
You should have received a copy of the GNU General Public License along with this program;
if not, write to the Free Software Foundation, Inc., 59 Temple Place, Suite 330, Boston, MA
Also add information on how to contact you by electronic and paper mail.
If the program is interactive, make it output a short notice like this when it starts in an interactive
mode:
Gnomovision version 69, Copyright (C) year name of author Gnomovision comes with
ABSOLUTELY NO WARRANTY; for details type `show w'. This is free software, and you are
welcome to redistribute it under certain conditions; type `show c' for details.
GNU General Public License 111

Open source licenses Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
The hypothetical commands `show w' and `show c' should show the appropriate parts of the General
Public License. Of course, the commands you use may be called something other than `show w' and
`show c'; they could even be mouse-clicks or menu items--whatever suits your program.
You should also get your employer (if you work as a programmer) or your school, if any, to sign a
"copyright disclaimer" for the program, if necessary. Here is a sample; alter the names:
Yoyodyne, Inc., hereby disclaims all copyright interest in the program `Gnomovision' (which
makes passes at compilers) written by James Hacker.
signature of Ty Coon, 1 April 1989
Ty Coon, President of Vice
This General Public License does not permit incorporating your program into proprietary
programs. If your program is a subroutine library, you may consider it more useful to permit
linking proprietary applications with the library. If this is what you want to do, use the GNU
Library General Public License instead of this License.
112 GNU General Public License

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Open source licenses 3-9000-759 Rev D June 2014
Version 3, 29 June 2007
Copyright © 2007 Free Software Foundation, Inc.
Everyone is permitted to copy and distribute verbatim copies of this license document, but
changing it is not allowed.
This version of the GNU Lesser General Public License incorporates the terms and conditions of
version 3 of the GNU General Public License, supplemented by the additional permissions listed
below.
As used herein, "this License" refers to version 3 of the GNU Lesser General Public License, and
the "GNU GPL" refers to version 3 of the GNU General Public License.
"The Library" refers to a covered work governed by this License, other than an Application or a
Combined Work as defined below.
An "Application" is any work that makes use of an interface provided by the Library, but which is
not otherwise based on the Library. Defining a subclass of a class defined by the Library is
deemed a mode of using an interface provided by the Library.
A "Combined Work" is a work produced by combining or linking an Application with the Library.
The particular version of the Library with which the Combined Work was made is also called the
"Linked Version".
The "Minimal Corresponding Source" for a Combined Work means the Corresponding Source for
the Combined Work, excluding any source code for portions of the Combined Work that,
considered in isolation, are based on the Application, and not on the Linked Version.
GNU Lesser General Public License 113

Open source licenses Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
The "Corresponding Application Code" for a Combined Work means the object code and/or
source code for the Application, including any data and utility programs needed for reproducing
the Combined Work from the Application, but excluding the System Libraries of the Combined
Work.
You may convey a covered work under sections 3 and 4 of this License without being bound by
section 3 of the GNU GPL.
If you modify a copy of the Library, and, in your modifications, a facility refers to a function or
data to be supplied by an Application that uses the facility (other than as an argument passed
when the facility is invoked), then you may convey a copy of the modified version:
"a) under this License, provided that you make a good faith effort to ensure that, in the event
an Application does not supply the function or data, the facility still operates, and performs
whatever part of its purpose remains meaningful, or
"b) under the GNU GPL, with none of the additional permissions of this License applicable to
that copy.
The object code form of an Application may incorporate material from a header file that is part of
the Library. You may convey such object code under terms of your choice, provided that, if the
incorporated material is not limited to numerical parameters, data structure layouts and
accessors, or small macros, inline functions and templates (ten or fewer lines in length), you do
both of the following:
a) Give prominent notice with each copy of the object code that the Library is used in it and
that the Library and its use are covered by this License.
b) Accompany the object code with a copy of the GNU GPL and this license document.
114 GNU Lesser General Public License

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Open source licenses 3-9000-759 Rev D June 2014
You may convey a Combined Work under terms of your choice that, taken together, effectively
do not restrict modification of the portions of the Library contained in the Combined Work and
reverse engineering for debugging such modifications, if you also do each of the following:
a) Give prominent notice with each copy of the Combined Work that the Library is used in it
and that the Library and its use are covered by this License.
b) Accompany the Combined Work with a copy of the GNU GPL and this license document.
c) For a Combined Work that displays copyright notices during execution, include the
copyright notice for the Library among these notices, as well as a reference directing the
user to the copies of the GNU GPL and this license document.
d) Do one of the following:
e) Provide Installation Information, but only if you would otherwise be required to provide such information under section 6 of the GNU GPL, and only to the extent that such information is necessary to install and execute a modified version of the Combined Work produced by recombining or relinking the Application with a modified version of the Linked Version. (If you use option 4d0, the Installation Information must accompany the Minimal Corresponding Source and Corresponding Application Code. If you use option 4d1, you must provide the Installation Information in the manner specified by section 6 of the GNU GPL for conveying Corresponding Source.)
GNU Lesser General Public License 115

Open source licenses Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
You may place library facilities that are a work based on the Library side by side in a single
library together with other library facilities that are not Applications and are not covered by this
License, and convey such a combined library under terms of your choice, if you do both of the
following:
a) Accompany the combined library with a copy of the same work based on the Library,
uncombined with any other library facilities, conveyed under the terms of this License.
b) Give prominent notice with the combined library that part of it is a work based on the
Library, and explaining where to find the accompanying uncombined form of the same
work.
The Free Software Foundation may publish revised and/or new versions of the GNU Lesser
General Public License from time to time. Such new versions will be similar in spirit to the
present version, but may differ in detail to address new problems or concerns.
Each version is given a distinguishing version number. If the Library as you received it specifies
that a certain numbered version of the GNU Lesser General Public License "or any later version"
applies to it, you have the option of following the terms and conditions either of that published
version or of any later version published by the Free Software Foundation. If the Library as you
received it does not specify a version number of the GNU Lesser General Public License, you may
choose any version of the GNU Lesser General Public License ever published by the Free
Software Foundation.
If the Library as you received it specifies that a proxy can decide whether future versions of the
GNU Lesser General Public License shall apply, that proxy's public statement of acceptance of
any version is permanent authorization for you to choose that version for the Library.
116 GNU Lesser General Public License

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Open source licenses 3-9000-759 Rev D June 2014
For more details about the Open SourceTM BSD license or the Open Source Initiative, follow the link below:
http://www.opensource.org/licenses/bsd-license.php
Copyright (c) ,
All rights reserved.
permitted provided that the following conditions are met:
and the following disclaimer.
conditions and the following disclaimer in the documentation and/or other materials provided with the distribution.
promote products derived from this software without specific prior written permission.
THIS SOFTWARE IS PROVIDED BY THE COPYRIGHT HOLDERS AND CONTRIBUTORS "AS IS" AND
ANY EXPRESS OR IMPLIED WARRANTIES, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED
WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE
DISCLAIMED. IN NO EVENT SHALL THE COPYRIGHT OWNER OR CONTRIBUTORS BE LIABLE FOR
ANY DIRECT, INDIRECT, INCIDENTAL, SPECIAL, EXEMPLARY, OR CONSEQUENTIAL DAMAGES
(INCLUDING, BUT NOT LIMITED TO, PROCUREMENT OF SUBSTITUTE GOODS OR SERVICES;
LOSS OF USE, DATA, OR PROFITS; OR BUSINESS INTERRUPTION) HOWEVER CAUSED AND ON
ANY THEORY OF LIABILITY, WHETHER IN CONTRACT, STRICT LIABILITY, OR TORT (INCLUDING
NEGLIGENCE OR OTHERWISE) ARISING IN ANY WAY OUT OF THE USE OF THIS SOFTWARE,
EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGE.
BSD Open Source License 117

Open source licenses Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
For more details about the Open SourceTM MIT license or the Open Source Initiative follow the link below:
http://www.opensource.org/licenses/mit-license.php
Copyright (c)
Permission is hereby granted, free of charge, to any person obtaining a copy of this software and
associated documentation files (the "Software"), to deal in the Software without restriction,
including without limitation the rights to use, copy, modify, merge, publish, distribute,
sublicense, and/or sell copies of the Software, and to permit persons to whom the Software is
furnished to do so, subject to the following conditions:
The above copyright notice and this permission notice shall be included in all copies or
substantial portions of the Software.
THE SOFTWARE IS PROVIDED "AS IS", WITHOUT WARRANTY OF ANY KIND, EXPRESS OR
IMPLIED, INCLUDING BUT NOT LIMITED TO THE WARRANTIES OF MERCHANTABILITY, FITNESS
FOR A PARTICULAR PURPOSE AND NONINFRINGEMENT. IN NO EVENT SHALL THE AUTHORS OR
COPYRIGHT HOLDERS BE LIABLE FOR ANY CLAIM, DAMAGES OR OTHER LIABILITY, WHETHER IN
AN ACTION OF CONTRACT, TORT OR OTHERWISE, ARISING FROM, OUT OF OR IN CONNECTION
WITH THE SOFTWARE OR THE USE OR OTHER DEALINGS IN THE SOFTWARE.
118 M.I.T License
The MIT License

Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Open source licenses 3-9000-759 Rev D June 2014
M.I.T License 119

Open source licenses Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual June 2014 3-9000-759 Rev D
Copyright (C) 1995-2005 Jean-loup Gailly and Mark Adler
This software is provided 'as-is', without any express or implied warranty. In no event will the
authors be held liable for any damages arising from the use of this software.
Permission is granted to anyone to use this software for any purpose, including commercial
applications, and to alter it and redistribute it freely, subject to the following restrictions:
the original software. If you use this software in a product, an acknowledgment in the product
documentation would be appreciated but is not required.
being the original software.
120 Zlib License





Daniel Measurement and Control, Inc. 11100 Brittmoore Park Drive Houston, TX 77041 USA
This product is a core component of the PlantWeb digital plant architecture.
T+1 713-467-6000 F+1 713-827-4805 USA Toll Free 1 888-356-9001 Daniel Measurement Services, Inc. T +1 713-827-6413 www.Daniel.com
Europe: Stirling, Scotland, UK T + 44-1786-433400 Middle East: Africa: Dubai, UAE T +971-4-811-8100 Asia Pacific: Singapore T +65-677-8211
© 2014 Daniel Measurement and Control, Inc., All rights reserved
The Emerson logo is a trademark and service mark of Emerson Electric Co. Daniel Measurement and Control, Inc. (Daniel) are Emerson Process Management business units. The Daniel name and logo are trademarks of Daniel Industries, Inc. The Senior, Junior, Simplex, MeterLink SeniorSonic, JuniorSonic, and DanPac are trademarks of Daniel Industries, Inc. All other trademarks are property of their respective owners.