T A B L E O F C O N T E N T S
1 Introduction
2 Warnings and Cautions
3 Specifications
4 Controls, Connectors and Visual Indicators
5 Operating Procedure
6 Cleaning and Disinfection
7 Accessories
8 Preventive Maintenance and Troubleshooting Guide
9 Warranty and Service

The Emerson CoughAssist™ Mechanical In-Exsufflator (MI-E) assists patients in
clearing retained bronchopulmonary secretions by gradually applying a positive
pressure to the airway, then rapidly shifting to a negative pressure. This rapid
shift in pressure, via a facemask, mouthpiece or an endotracheal or tracheostomy
tube, produces a high expiratory flow rate from the lungs, simulating a cough, a
technique referred to as “mechanical insufflation-exsufflation.” The automatic
CoughAssist MI-E (CA-3000, CA-3200) has timing mechanisms to automate the
inspiratory and expiratory cycles as well as a manual control. The manual
CoughAssist MI-E (CM-3000, CM-3200) uses a manually operated valve to shift
from positive to negative pressure and back.
Those who might benefit from the use of the CoughAssist MI-E include any patient
with an ineffective cough due to muscular dystrophy, myasthenia gravis, poliomyelitis,
or other neurologic disorder with some paralysis of the respiratory muscles, such
as spinal cord injury. It may also be used to treat ineffective cough due to other
bronchopulmonary diseases, such as emphysema, cystic fibrosis and bronchiectasis.
It is effective for both trached and non-invasively ventilated patients.
due to reduced peak cough expiratory flow (less than 2 to 3 liters per second),
resulting from high spinal cord injuries, neuro-muscular deficits or severe
fatigue associated with intrinsic lung disease, is a candidate for this device.
known susceptibility to pneumothorax or pnuemo-mediastinum, or known to
have had any recent barotrauma, should be carefully considered before use.

SECTION 2 WA R N I N G S A N D C A U T I O N S
Per IEC 60601-1, Medical Electrical Equipment, General Requirements for Safety, the CoughAssist MI-E is classified as follows:
Class 1 Equipment: Equipment in which protection against electric shock does not rely on basic insulation only, but includes a grounding pin on the power cord. For ground reliability always plug the power cord into an AC grounded outlet.
Type BF Equipment: Type B piece of equipment with an F-Type applied part.
A Type B piece of equipment is one that provides a particular degree of protection against electric shock, particularly regarding allowable leakage current and reliability of the protective earth connection (grounding).
F-Type applied part is one that extends from the patient into the equipment and is isolated from all other parts of the equipment.
Water Ingress: This device does not have protection against ingress of water.
Disinfection: With the exception of the patient circuit, this device can be disinfec- ted using 70% isopropyl alcohol or equivalent. (See Section 6: Cleaning and Disinfection.)
Flammable Anesthetics: This device is not suitable for use in the presence of a flammable anesthetic mixture with air, or in the presence of a flammable anesthetic mixture with oxygen or nitrous oxide.
Intermittent Operation: This device is designed for Intermittent Operation Only and not for continuous use. The device should not be cycled continuously for more than 5 minutes. After such time, the unit should either be turned off or left idling with the blower on for at least 5 minutes.

D E F I N I T I O N S
Throughout this guide the following definitions apply:
V Patients known to have cardiac instability should be monitored for pulse and oxygen saturation very closely.
V Soreness and/or pain in the chest from a pulled muscle may occur in patients using the CoughAssist MI-E for the first time if the positive pressure used exceeds pressures, which the patient normally receives during Positive Pressure Therapy.1
Such patients should start at a lower positive pressure during treatment, and gradually (over several days, or as tolerated) increase the positive pressure used.
V Do not use in the presence of flammable anesthetics.
V Connection should be made to a grounded outlet only.
V Do not place or store the device where it can fall or be pulled into a tub or sink.
V If the device comes into contact with water, unplug the unit.
V Never operate the CoughAssist MI-E if it has a damaged cord or plug, is not work- ing properly, or has been dropped, damaged or immersed in water.
1 Positive Pressure Therapy includes the use of a volume ventilator, nasal or mask ventilation or CPAP (Continuous Positive Airway Pressure), or IPPB (Intermittent Positive Pressure Breathing).

WA R N I N G S A N D C A U T I O N S
V Replace fuses only with ones having the same ratings for blow characteristics, current and voltage.
V Do not remove the cover; there are no serviceable parts inside the unit. Refer all service to authorized personnel.
V Always check time and pressure settings before each treatment.

S P E C I F I C A T I O N S
Maximum Positive Pressure:
60 cm H2O (44 mm Hg)
Maximum Negative Pressure:
60 cm H2O (44 mm Hg)
Standards:
Conforms to UL STD 2601-1, certified to CAN/CSA - STD C22.2 No. 601.1-M90, 0413
Maximum Inhalation Flow:
3.3 liters/second with inhale flow set to minimum; if set to maximum inhalation, the flow is the same as the exhalation flow
Maximum Exhalation Flow Capacity:
10 liters/second; actual flow depends on maximum pressure and on airway resistance
Pressure Gauge:
-70 to 0 to +70 cm H2O; accuracy +/-2% of full scale
Mode of Operation:
Inhalation, Exhalation, Pause Times:
“Off” Position:
Yes - connects to ambient
Blower Type:
Two-stage centrifugal blower with AC/DC universal motor
Input Voltage:
Input Power:
Dimensions:
292mm x 279mm x 419mm (HxWxD) (11.5” x 11” x 16.5”)
Weight:
Temperature Specifications:
Operating temperature range: +10 to +40°C (+50 to +104°F) Storage temperature range: -20 to +50°C (-4 to +122°F)
Humidity Specifications:

CA-3000, CA-3200 F R O N T PA N E L C O N T R O L S
exhale pressure.
Use only if the gauge does not return to “O” when the unit is turned off (see Section 8).
o

CM-3000, CM-3200 F R O N T PA N E L C O N T R O L S
exhale pressure.
Use only if the gauge does not return to “0” when the unit is turned off (see Section 8).
o


B A C K PA N E L ( A L L M O D E L S)



O P E R A T I N G P R O C E D U R E
I N I T I A L S E T- U P
a. Attach the bacterial/viral filter to the patient port on the front panel.
b. Attach the 3-foot 22mm ID smooth bore breathing hose to the bacterial/viral filter.
c. Attach the appropriate patient interface to the breathing hose. Patient interface options include a facemask and adapter, mouthpiece, lip seal or tracheostomy tube adapter. (A facemask and adapter are included with each unit.)
2 CoughAssist MI-E meets IEC 60601-1-2 for Electromagnetic Interference and Immunity, Medical Electrical Equipment, Part 1, General Requirements for Safety, 2. Collateral Standard: Electromagnetic Compatibility Requirements and Tests, which defines standards for both electromagnetic emissions and immunity.

P R E S S U R E A D J U S T M E N T
Each individual patient may require special settings for the maximum positive (inhalation) and negative (exhalation) pressures. For a patient using this device for the first time, it is advisable to begin with lower pressures, such as 10-15 cm H2O positive and negative, to familiarize the patient with the feel of mechanical insufflation-exsufflation. During subsequent treatments, pressures can be increased as necessary to achieve adequate secretion clearance. See WARNING (page 4).

O P E R A T I N G P R O C E D U R E
T I M I N G A D J U S T M E N T (CA-3000, CA-3200 M O D E L S O N LY)
If the automatic feature of the CoughAssist MI-E is to be used, adjust the times as follows:
Treatment usually consists of 4 or 5 coughing cycles in succession. The patient is then allowed to rest for 20 to 30 seconds, which helps avoid hyperventilation. The cycles can then be repeated 4 to 6 times for a full treatment.
Manual Operation (All models):

O P E R A T I N G P R O C E D U R E
Automatic Operation (CA-3000, CA-3200 models only):
It is recommended that the CoughAssist MI-E be periodically tested to ensure that the cycling valve returns to the neutral, or pause, position after either the inhale or exhale phase. To determine this, follow these steps:

C L E A N I N G & D I S I N F E C T I O N
PAT I E N T C I R C U I T
Institutional (hospital) Use:
Home (individual) Use:
E X T E R N A L H O U S I N G
The exterior of the pump and housing may be washed with a mild detergent and water, or with a bactericidal cleaning solution such as 70% isopropyl alcohol.

The following replacement accessories may be obtained from J.H. Emerson Co.

CoughAssist MI-E Patient Circuit
The CoughAssist MI-E Patient Circuit consists of one 3-foot long flexible smooth bore tube, a bacterial/viral filter, an adult face- mask and an adapter.
Part No. 325-9217

3-foot long flexible smooth bore tubing with 22mm interior diameter.
Part No. 732-1136

Bacterial/Viral filter.
Part No. 740-1006
Part No. 740-1008

Facemask and Adapter, 22mm outside diameter x 22mm outside diameter.
Part No. 740-1007
Breathing Hose · Bacterial/Viral Filter · Facemask and Adapter

P R E V E N T I V E M A I N T E N A N C E &
T R O U B L E S H O O T I N G G U I D E
P R E V E N T I V E M A I N T E N A N C E
This device has been designed to provide virtually maintenance-free operation for extended periods of time. Sharp blows to the unit or dropping the unit is to be avoided. No routine maintenance is required.
G E N E R A L S U G G E S T I O N S
Technical Information: J.H. Emerson Co. will make available on request a list of all repairable exterior parts with descriptions. Interior schematics and circuit
diagrams will be made available to qualified technical personnel only.

T R O U B L E S H O O T I N G G U I D E
a. Disconnect the unit from any power outlet and disconnect the power cord from the receptacle on the rear of the unit (see Back Panel diagram on page 9).
b. Locate the access door on the receptacle labeled with the symbol: . Open the access door by prying the latch at the top with a small screw driver or fingernail. Pivot the door down to reveal the two fuse holders.

c. Press each of the two spring clips to the side (the left one to the left, the right one to the right) and slide both fuse holders out of the receptacle.

d. Inspect both fuses and replace, if necessary, with fuses with equivalent ratings, as shown on the rear Product Label Description (page 9). To replace a fuse, slide the damaged one out of the holder and slide the new fuse in its place.
e. Reinsert each fuse and holder into the receptacle, close the access door and reconnect the power cord. 17

W A R R A N T Y & S E R V I C E
The CoughAssist MI-E is warranted to be free of defects in material or workmanship for one year.
For questions about this product please con- tact Customer Service at:
J.H. Emerson Co.
22 Cottage Park Avenue
REPRESENTATIVE
FOR THE EUROPEAN