Agilent PN 8510-16 Controlling T est Port Output Power Flatness Product Note UNCORRECTED POWER CORRECTED POWER OUTPUT POWER Agilent 8510C Network Analyzer.
2 Introduction Designers and manufacturers of active devices oft en need to control the pow er lev el at the t est port 1 of t heir pow er -sensitive devices, but f ind dif- ficulty in o vercoming inser tion losses. The insertion losses occur as a result of connecting components in the measurement pat h between the source and the DUT .
3 System Configuration The basic 8510C measurement system conf igura- tion needed for test por t f latness correction is sho wn in Figure 1. A sing le channel pow er meter such as the A gilent 437B, E4418A, or E4418B is required, along with a compatible sensor , for per- forming t he po wer f latness calibration.
4 Flatness-Correction Operation The basic setup procedure t o obtain f lat t est port output po wer is illustrated in Table 1. To simplify the execution of t he procedure, t he front panel hardk eys are enclosed in [brackets]. The sof tkeys are enclosed in {braces}.
5 T able 1. Flatness-correction calibration procedure 8510C Keystrokes Description Set up the power meter (see T able 2) For proper operation, the E4418B must be set up before initiating the flatness-correction routine. See T able 2 for specific instructions.
6 T able 2. Agilent E4418B power meter setup E4418B Keystrokes Description Preset and zero the power meter [PRESET/LOCAL] Return the power meter to a known state. {CONFIRM} [ZERO/CAL] {ZERO} Set up the 437B emulation on the power meter [SYSTEM/INPUTS] The 437B command set must be activated in order to perform the power flatness calibration.
7 The follo wing er ror messages may occur while attempting t o enable the f latness cor rection and/or reducing the test por t pow er level: • If f latness cor rection is enabled before reducing the test por t pow er level, IF Overload or Source 1 W arning—RF Unleveled ma y be display ed on the analyzer as the source becomes unlev eled.
8 V erifying a flatness-correction calibration T o verify t he f latness-correction calibration, t he po wer sensor should be reconnected t o the t est port t o measure the t est por t po wer at individual frequencies. Since the measurement system is calibrated in 50W , inaccuracies will occur when a device is not well mat ched.
9 Using test port 1 calibrations on test port 2 In most cases, port 1 will be t he input por t of t he DUT . When por t 2 must be used as the input por t to t he device, t he user may choose to use a por t 1 f latness-cor rection calibration on por t 2 since the port 1 and por t 2 signal pat hs are symmetrical.
10 Practical Application Examples Absolute output power measurements with flatness correction One benefit of t he f latness-correction capability is the ability t o measure the absolute output pow er of active devices.
11 Amplifier measurement example Step-by-step instr uctions for setting up and apply- ing f latness cor rections for t he measurement of an amplifier are sho wn in Table 7 . Gain, absolute output po wer , and gain compression measurements procedures are cov ered.
Agilent T echnologies’ T est and Measurement Support, Services, and Assistance Agilent Technologies aims to maximize the value you receive, while minimizing your risk and problems. W e strive to ensure that you get the test and measurement capabilities you paid for and obt ain the suppor t you need.
An important point after buying a device Agilent Technologies PN 8510-16 (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Agilent Technologies PN 8510-16 yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Agilent Technologies PN 8510-16 - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Agilent Technologies PN 8510-16 you will learn all the available features of the product, as well as information on its operation. The information that you get Agilent Technologies PN 8510-16 will certainly help you make a decision on the purchase.
If you already are a holder of Agilent Technologies PN 8510-16, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Agilent Technologies PN 8510-16.
However, one of the most important roles played by the user manual is to help in solving problems with Agilent Technologies PN 8510-16. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Agilent Technologies PN 8510-16 along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center