Revision 2.7
FuturePlus Systems is a trademark of FuturePlus Systems Corporation Copyright 1998 FuturePlus Systems Corporation

PRODUCT WARRANTY 5
Limitation of warranty 5 Exclusive Remedies 6
How to Use This Manual 7
Duplicating the Master Diskette 9
Minimum Equipment Required 10
Signal Naming Conventions 10
The USB Analysis Probe Switches and Diodes 10 The Slow Speed Only Switch and LED 10 USB Busy LED 10 Ready LED 10 VCC LED 10
Connecting to the USB Analysis Probe 11
Installing the USB Analysis Probe 11
Setting up the Analyzers installed in the 16500 mainframe. 12
Setting up the 167xx Analyzer 12
Setting up the 1680/90/900 Analyzer 13 1680/90/900 licensing 13
Loading 1680/90/900 configuration files 13
1680/90/900, 167xx Configuration files 15
The Format Menu 18

The STAT variable 19 The ADDR and DATA variables 19 The PID variable 19 The CTLCMD variable 21
Bad EOP Detection 22
Low Speed/High Speed detection 23
Installation Quick Reference 25
The State Display 26
Analysis Probe related error messages 27
USB CRC and Serial Bit Errors 27
Installation Quick Reference 29
POD 3 Description 30 End of Packet State Machine 31 Load Byte Count State Machine 32
USB Analysis Probe Test Points 33 POD 4 Description 34
The Waveform Display 34
Characteristics 35 Standards Supported 35 Power Requirements 35 Logic Analyzer Required 35 Number of Probes Used 35 Supported speeds 35 Signal loading 35 Operations 35 Environmental Temperature 36 Altitude 36 Humidity 36 Testing and Troubleshooting 36 Servicing 36
Acquiring Data 26 · Acquiring Data 30 · Signal Connections 36

For Technical Support:
36 Olde English Road
FAX: 603-471-2738
On the web http://www.futureplus.com
For Sales and Marketing Support:
FAX: 719-278-9586
On the web http://www.futureplus.com
FuturePlus Systems has technical sales representatives in several major countries. For an up to date listing please see http://www.futureplus.com/contact.html
Agilent Technologies is also an authorized reseller of many FuturePlus products. Contact any Agilent Technologies sales office for details.
FuturePlus Systems Corporation · Bedford NH 03110 · FuturePlus Systems Corporation

This FuturePlus Systems product has a warranty against defects in material and workmanship for a period of 1 year from the date of shipment. During the warranty period, FuturePlus Systems will, at its option, either replace or repair products proven to be defective. For warranty service or repair, this product must be returned to the factory.
For products returned to FuturePlus Systems for warranty service, the Buyer shall prepay shipping charges to FuturePlus Systems and FuturePlus Systems shall pay shipping charges to return the product to the Buyer. However, the Buyer shall pay all shipping charges, duties, and taxes for products returned to FuturePlus Systems from another country.
FuturePlus Systems warrants that its software and hardware designated by FuturePlus Systems for use with an instrument will execute its programming instructions when properly installed on that instrument. FuturePlus Systems does not warrant that the operation of the hardware or software will be uninterrupted or error-free.
The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by the Buyer, Buyer-supplied software or interfacing, unauthorized modification or misuse, operation outside of the environmental specifications for the

product, or improper site preparation or maintenance. NO OTHER WARRANTY IS EXPRESSED OR IMPLIED. FUTUREPLUS SYSTEMS SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND EXCLUSIVE REMEDIES. FUTUREPLUS SYSTEMS SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER BASED ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY.
Assistance Product maintenance agreements and other customer assistance agreements are available for FuturePlus Systems products. For assistance, contact the factory.
Exclusive Remedies

The USB Analysis Probe module provides a complete interface between any point on the USB and an Agilent Logic Analyzer. The Analysis Probe interface receives the USB serial bit stream and converts it to an understandable parallel bus that is then acquired by the logic analyzer.
The USB Analysis Probe is a passive bus monitor which does not assert any signals on the USB . The USB signals are received with a USB compliant transceiver. The power for the Analysis Probe logic and transceiver is taken from the logic analyzer. Thus the Analysis Probe does not draw power from the USB wire.
The USB Analysis Probe software configures all the logic analyzer menus for compatibility with the USB Analysis Probe. When the configuration file is loaded and the analyzer is set to STATE mode, an inverse assembler is also loaded which decodes USB packets into easy to read mnemonics.
The USB Analysis Probe also provides access to the actual USB wire for both digital and analog measurements. D+, D-, GND, USB Power and the transceiver outputs of VP, VM, and RCV are also provided on the test points.
This manual is organized to help you quickly find the information you need.


This chapter introduces you to the USB Analysis Probe and lists the minimum equipment required and accessories supplied for Universal Serial Bus analysis. This chapter also contains information that is common to both state and timing analysis.
Before you use the USB Analysis Probe software, make a duplicate copy of the master diskette. Then store the master diskette and use the back-up copy to configure your logic analyzer. This will help prevent the possibility of losing or destroying the original files in the event the diskette wears out, is damaged, or a file is accidentally deleted.
To make a duplicate copy, use the Duplicate diskette operation in the disk menu of your logic analyzer. For more information, refer to the reference manual for your logic analyzer.
The USB Analysis Probe product consists of the following accessories:

The minimum equipment required for analysis of a Universal Serial Bus consists of the following equipment:
Revisions This manual applies to Hardware revisions ACE or later and software revision 3.0 or later.
This operating manual uses the same signal notation as the UNIVERSAL SERIAL BUS SPECIFICATION - REVISION 1.0.
The USB Analysis Probe contains one switch and four LEDs
The Slow Speed Only Switch and LED
The slow speed switch should be in the ON position (LED ON) when the USB Analysis Probe is on a slow speed only wire or slow speed only segment of the USB. When attached to a high speed segment or mixed segment (slow speed and high speed traffic) the switch should be in the OFF position (LED OFF).
The USB Busy LED is lit when the USB Analysis Probe is actively receiving packets from the USB wire. The LED may appear dim if the traffic load is light.
Ready LED The Ready LED indicates that the on board serial ROM has successfully loaded the FPGA and that the USB Analysis Probe is ready for USB analysis.
VCC LED The VCC LED on the USB Analysis Probe indicates that power is applied from the logic analyzer to the USB Analysis Probe logic. For correct operation of the USB Analysis Probe the VCC LED must be lit.
USB Busy LED

The following explains how to connect the logic analyzer to the USB Analysis Probe for either state or timing analysis:
The USB Analysis Probe can be installed in any segment of the Universal Serial Bus. However, to view all packets from the Host the USB Analysis Probe must be placed between the Host and the USB Hub or USB device. The following steps explain how to install the USB Analysis Probe onto the Universal Serial Bus.

Setting up the Analyzers installed in the 16500 mainframe.
The logic analyzer can be configured for USB analysis by loading the USB configuration file. Loading this file will load the Universal Serial Bus inverse assembler and configure your logic analyzer for USB analysis.
The 16600/16700 requires a special install procedure to install the FS4100 software. To accomplish this, insert the diskette labeled 16700/702 Installation disk for the FS4100 into the 16700/702 diskette drive. From the SYSTEM ADMINISTRATION TOOLS select INSTALL under SOFTWARE. From the SOFTWARE INSTALL screen select the FLEXIBLE DISK and APPLY. The package FS4100 will now appear. Select it and then select INSTALL. This procedure does not need to be repeated. It only needs to be done the first time the USB Analysis Probe is used.
When this has completed restart the logic analysis session and either invoke the Setup Assistant from the logic analyzer screen or load the appropriate configuration file from the Configs/FuturePlus/FS4100 directory. The Setup Assistant will guide you in configuring the logic analyzer. Select FuturePlus from the list on the left of the Setup Assistant screen and then USB from the list that then appears. If you prefer to load the

configuration file yourself, see table below for a list of analyzers and corresponding configuration files.
The 1680/90/900 Analyzer is a PC based application that requires a PC running the Windows OS or a 16900 frame.
Setting up the 1680/90/900 Analyzer Before installing the protocol decoder for the USB protocol on a PC you must install the Agilent logic analyzer software. Once the Agilent logic analyzer software is installed, you can install the FS4100 protocol decoder by placing the CD-ROM disk into the CD-ROM drive of the target computer or Analyzer and executing the .exe setup program that is contained on the disk. The .exe setup file can be executed from within the File Explorer PC Utility. You must navigate to the .exe file on the CD-ROM disk and then double click the .exe file name from within the File Explorer navigation panel.
The installation procedure does not need to be repeated. It only needs to be done the first time the Analysis Probe Adapter is used.
The USB Inverse Assembler is a licensed product that is locked to a single hard drive. The licensing process is performed by Agilent. There are instructions on this process on the SW Entitlement certificate provided with this product.
When the software has been licensed you should be ready to load a configuration file. You can access the configuration files by clicking on the folder that was placed on the desktop. When you click on the folder it should open up to display all the configuration files to choose from. If you put your mouse cursor on the name of the file a description will appear telling you what the setup consists of, once you choose the configuration file that is appropriate for your configuration the 16900 operating system should execute. The protocol decoder automatically loads when the configuration file is loaded. If the decoder does not load, you may load it by selecting tools from
1680/90/900 licensing

the menu bar at the top of the screen and select the decoder from the list.
After loading the configuration file of choice, go into the format specification of the configuration by choosing Setup from the menu bar and then selecting Bus/Signal in the drop down menu. When the format specification appears press Define Probes at the bottom of the screen. The Define Probes feature will describe how to hook the analyzer cards to the connections on the target. The following figure shows what the Define Probes screen looks like. The figure below may differ from your display; this is an example of how the display looks in general.


1680/90/900, 167xx Configuration files
167xx Analyzer 169xx Analyzer File Name
16555/6/7 CUSB41_2
16715/6/7, 1674x, 1675x CUSB41_3
1680/90 (State) 1680/90, 16750/1/2, 1691x (State) CUSB41_6
1680/90 (Timing) 1680/90, 16750/1/2, 1691x (Timing) CUSB41_7
Offline Analysis Data that is saved on a 167xx analyzer in fast binary format, or 16900 analyzer data saved as a *.ala file, can be imported into the 1680/90/900 environment for analysis. You can do offline analysis on a PC if you have the 1680/90/900 operating system installed on the PC, if you need this software please contact Agilent.
Offline analysis allows a user to be able to analyze a trace offline at a PC so it frees up the analyzer for another person to use the analyzer to capture data.
If you have already used the license that was included with your package on a 1680/90/900 analyzer and would like to have the offline analysis feature on a PC you may buy additional licenses, please contact FuturePlus sales department.
In order to view decoded data offline, after installing the 1680/90/900 operating system on a PC, you must install the FuturePlus software. Please follow the installation instructions for “Setting up 1680/90/900 analyzer”. Once the FuturePlus software has been installed and licensed follow these steps to import the data and view it.
From the desktop, double click on the Agilent logic analyzer icon. When the application comes up there

will be a series of questions, answer the first question asking which startup option to use, select Continue Offline. On the analyzer type question, select cancel. When the application comes all the way up you should have a blank screen with a menu bar and tool bar at the top.
For data from a 1680/90/900 analyzer, open the .ala file using the File, Open menu selections and browse to the desired .ala file.
For data from a 16700, choose File -> Import from the menu bar, after selecting import select “yes” when it asks if the system is ready to import 16700 data.

After clicking “next” you must browse for the fast binary data file you want to import. Once you have located the file and clicked start import, the data should appear in the listing.
After the data has been imported you must load the protocol decoder before you will see any decoding. To load the decoder select Tools from the menu bar, when the drop down menu appears select Inverse Assembler, then choose the name of the decoder for your particular product. The figure below is a general

picture; please choose the appropriate decoder for the trace you are working with.

After the decoder has loaded, select Preferences if required, from the overview screen and set the preferences to their correct value in order to decode the trace properly. This is a general requirement, some decoders do not have preferences, if this is the case then no preference setting is necessary.

The USB Analysis Probe diskette sets up the format menu as shown in the following table. This format is the same for both Timing and State Analysis. Pods 1 and 2 are for STATE analysis and POD 3 is useful for Timing analysis.

Label Pod 4 Pod 3 Pod 2 Pod 1

The STAT variable is used by the USB inverse assembler to decode USB transactions. It should not be changed or deleted from the format menu.
The ADDR and DATA variables
The ADDR variable is the address and endpoint as decoded from the USB serial bit stream. The DATA variable is the Data associated with a Data0 or Data1 packet.
For ease of triggering the address and endpoint seen in the token packets Setup, In and Out are latched by the USB Analysis Probe and held through the reporting of the data for the DATA0 and DATA1 packets.
The PID variable describes the PID and various errors detected by the USB Analysis Probe. This variable is most effectively used for triggering and are available to be used in combination with any other variable defined in the FORMAT menu.

The STAT variable · The PID variable


IN SBS ERROR 001010 IN PACKET WITH SERIAL BIT STUFFED


010100 START OF FRAME PACKET WITH CRC ERROR DETECTED
110111 DETECTED INCORRECT J STATE AS EOP SIGNALING AT
The CTLCMD variable decodes the Setup Data into the appropriate command. The CTLCMD variable


SET_CONFIGURATION
GET_INTERFACE
SET_INTERFACE
SYNCH_FRAME
COMBINATION MODE
The default mode set up by the software shipped with the Analysis Probe is the following configuration.

Bad EOP Detection If the FS4100 operating in full speed mode, detects a valid SE0 for at least 60ns but a valid J_STATE does not follow, it will do the following.
If the FS4100 operating in slow speed mode, detects a valid SE0 for at least 300ns but a valid J_STATE does not follow, it will do the following.
Standard Device Requests

A status of “POSSIBLE BAD EOP” should be treated as cautionary and the user would be advised to view the received signals in timing mode by doing the following.
The USB Analysis Probe can switch automatically from high speed to low speed upon the detection of the Preamble PID. The Analysis Probe then detects only low speed traffic. Upon the receipt of a low speed acknowledge or an internal time-out (greater than 16 low speed bit times). The Analysis Probe will switch back to high speed. The following explains in more detail the sequences the Analysis Probe looks for in making the decision whether to switch from high speed to low speed and back.
Sequence 1

If the Analysis Probe is placed on a low speed only segment the low speed only switch must be placed in the on position (LED lit). This is because on a low speed only segment the Host or Hub will not send a Preamble PID. NOTE: The low speed only switch need only be placed in the ON position if the Analysis Probe is placed on a LOW SPEED ONLY SEGMENT (no high speed traffic present).
If users are on a mixed high speed/low speed segment and suspect that low speed/high speed USB specification protocol is not being adhered to, the following steps should be taken.
Sequence 2

This chapter explains how to configure the USB Analysis Probe to perform state analysis on the Universal Serial Bus. The configuration software on the flexible diskette sets up the format specification menu of the logic analyzer for compatibility with the USB Analysis Probe. The next chapter explains how to configure the USB Analysis Probe to perform timing analysis.
The following procedure describes the major steps required to perform measurements with the USB Analysis Probe module.
The following explains how to connect the logic analyzer to the USB Analysis Probe for state analysis:
Logic Analyzer USB Analysis
Master POD 1 Header 1 State analysis (USB_ST)
POD 2 Header 2 State analysis (USB_ST)
Comment · Probe

Probe OUT connector can be connected down stream to either a Hub or device or it can be left unconnected.
READY LEDs on the USB Analysis Probe are lit. NOTE: Connect the USB Analysis Probe to the USB wire as described in step 2 prior to powering up the logic analyzer.
The logic analyzer is now ready for STATE analysis.
Acquiring Data Touch RUN and as soon as there is activity on the bus, the logic analyzer will begin to acquire data. The analyzer will continue to acquire data and will display the data when the analyzer memory is full, the trigger specification is TRUE or when you touch STOP.
The logic analyzer will flash “Slow or Missing Clock” if the USB Analysis Probe provided master clock signal is not being detected by the logic analyzer. This will occur if the USB is IDLE or in an extended suspend state. To accurately determine the state of the USB refer to the Timing analysis chapter in this manual.
Captured data is as shown in the following figure. The following figure displays the state listing after disassembly. The inverse assembler is constructed so the mnemonic output closely resembles the actual commands, status conditions, messages and phases specified in the Universal Serial Bus specification. Symbols on the PID variable have also been defined to help aid in analysis. The non-disassembled state listing displays USB mnemonics in addition to data. All data, address, endpoint and frame number fields are displayed in hex.


Analysis Probe related error messages
The following Analysis Probe related error messages are reported by the USB inverse assembler.
This error will be reported by the inverse assembler if the USB Analysis Probe hardware reports a status code that is undefined. If this error occurs please contact the FuturePlus Systems factory.
By design the USB Analysis Probe will detect and report CRC and serial bit stuffed errors. The PID variable chart details the CRC and serial bit stuff errors detected.
It is a protocol violation to omit the stuffed zero when the last 6 bits of a packet are ones although the packet data is correct and complete. The Analysis Probe may not indicate that a serial stuffed bit error occurred in this scenario.

The USB Analysis Probe will also detect invalid PIDS, bad PIDS and SOF tokens sent at slow speed. The Analysis Probe implements an error summary status called ANY ERROR, this status will be sent to the Analyzer if any of the above error conditions is detected by the Analysis Probe.

Pod 3 of the USB Analysis Probe contains signals that describe in more detail the actual state of the USB wire.
The following procedure describes the major steps required to perform timing analysis measurements with the USB Analysis Probe module.

Acquiring Data Touch RUN and the logic analyzer will begin to acquire data. The analyzer will continue to acquire data and will display the data when the analyzer memory is full, the trigger specification is TRUE or when you touch STOP.
The logic analyzer will flash “Waiting for Trigger” or “occurrences remaining in level x” where x is number of the unsatisfied trigger level if the trigger condition is not satisfied.
POD 3 Description Pod 3 was included in the USB Analysis Probe to give the user a detailed look at the USB wire. The USB Analysis Probe interface contains its own fully compliant USB serial interface engine (SIE). The state bits for the USB state machines for this SIE are available on POD 3. In addition the recovered clock, recovered data, start of frame, end of packet, end of sync and low speed/high speed detect signals are available. These signals are useful for :
POD 3 channel Signal Name Description
16 CLK12 Recovered Clock
15 MDATA Recovered Serial Data


14 SOFTIC Start of Frame. 1 millisecond timer generated from recovered start of frame
The end of packet state machine state encodings are as follows: End of Packet State Machine

SEO_FIRST 100 First sample of the single ended zero condition
SEO_SECOND 010 Second sample of the single ended zero condition
SEO_THIRD 110 Third or more sample of the single ended zero condition


EOPR_STATE 001 J state transition detected. Valid end of packet
101 First slow speed single ended zero condition detected
011 First slow speed single ended zero condition detected
111 Incorrect transition for single ended zero detected. NOTE: FEOPR will be generated in this case so that the next packet can be detected. Users can look for this state if they suspect a bad end of packet condition occurring on the USB.
The load byte count state machine is encoded as follows. Load Byte Count State Machine State Name State Encoding Description

PID 0001 Have received end of sync and currently receiving PID

DELAY_1 0010 State delayed by one clock tic
DELAY_2 0011 State delayed by two clock tics
SOF_OR_ADR 0100 Receiving address for token packet or Frame number
SOF_OR_EP 0101 Receiving endpoint for token packet or the remainder of the frame number for start of frame
CRC 0110 Receiving CRC data
PRE_DATA 0111 About to receive data
DATA 1000 Receiving data
D+, D-, VP, VM, RCV, USB Power and GND are available to be viewed with spare probes from the logic analyzer or a scope probe.
The format menu has been configured to include POD4. Pod 4 allows the user by means of attaching a flying led set to the stake pins to view the signals that

the transceiver is generating in response to the USB traffic.
POD 4 channel Signal Name Description
0 RCV Differential data signal from the USB transceiver
1 VM VM Signal from the USB transceiver
2 VP VP Signal from the USB transceiver.
Captured data is displayed as shown in the following figure.

POD 4 Description

This chapter provides additional reference information including the characteristics and signal connections for the USB Analysis Probe module.
Characteristics The following operating characteristics are not specifications, but are typical operating characteristics for the USB Analysis Probe.
Standards Supported The Universal Serial Bus Specification Revision 1.0
Power Requirements The USB Analysis Probe does not draw power from the USB wire. All power is provided by the logic analyzer.
Logic Analyzer Required 166x, 167x, 1655x
USB State Analysis - 2 PODS Number of Probes Used
USB Timing Analysis - 3 PODS
Supported speeds The USB Analysis Probe supports both full speed and low speed operation.
Signal loading The USB Analysis Probe presents one USB transceiver load onto the USB. The Analysis Probe IN and OUT connectors (D+, D-, GND and VCC) are daisy chained so that the USB traffic is not interrupted.
All signal etch length, trace velocity and impedance is within specification.
Operations All Universal Serial Bus operations supported.

Non operating:-40 to +75 degrees C (-40 to +167 degrees F)
Up to 90% non condensing. Avoid sudden, extreme temperature changes which would cause condensation on the Analysis Probe module.
There are no automatic performance tests or adjustments for the USB Analysis Probe module. If a failure is suspected in the USB Analysis Probe contact the factory or your FuturePlus Systems authorized distributor.
Servicing The repair strategy for the USB Analysis Probe is replacement. However, if parts of the USB Analysis Probe module are damaged or lost contact the factory for a list of replacement parts.
Signal Connections The USB Analysis Probe module monitors signals for both state and timing analysis. The below figure displays how the cable headers are numbered.
39 37 35 33 31 29 27 25 23 21 19 17 15 13 11 9 7 5 3 1
40 38 36 34 32 30 28 26 24 22 20 18 16 14 12 10 8 6 4 2
The following tables list the USB Analysis Probe cable headers and the corresponding Universal Serial Bus signals.
Environmental Temperature · Humidity · Testing and Troubleshooting






Pages referenced in this manual — manufacturer, support and downloads.
Owners and our team usually reply within a day — and your question helps the next person too.
Be the first to ask — other owners or our team will answer.