Hard Disk Drive Kit
User’s Guide

Agilent Part No. E5070-90119
June 2005 · Printed in Japan

The information contained in this document is subject to change without notice.
This document contains proprietary information that is protected by copyright.All rights are reserved. No part of this document may be photocopied, reproduced, or translated to another language without the prior written consent of the Agilent Technologies.
© Copyright 2005 Agilent Technologies
The manual’s printing date and part number indicate its current edition. The printing date changes when a new edition is printed. (Minor corrections and updates that are incorporated at reprint do not cause the date to change.) The manual part number changes when extensive technical changes are incorporated.
May 2005 First Edition
June 2005 Second Edition
Sample (bold) Boldface type is used when a term is defined or emphasis.
Sample (Italic) Italic type is used for emphasis.
key Indicates a hardkey (key on the front panel or external keyboard) labeled “Sample.” “key” may be omitted.
Sample menu/button/box Indicates a menu/button/box on the screen labeled “Sample” which can be selected/executed by clicking. “menu,” “button,” or “box” may be omitted.
Sample block/toolbar Indicates a block (group of hardkeys) or a toolbar (setup toolbar) labeled “Sample.”
Sample 1 - Sample 2 - Sample 3 Indicates a sequential operation of Sample 1, Sample 2, and Sample 3 (menu, button, or box).
“-” may be omitted.

This chapter describes the Kit, contents of the kit, and required tools to install the Option 018 into the Agilent E5070B/E5071B.


This chapter describes the procedure to install the option 018 into the Agilent E5070B/E5071B.

This chapter describes the procedure to install Option 018 into the E5070B/E5071B. The procedure consists of the following items.
Follow the steps below to remove a removable hard disk.
the cable’s connector by pressing the locking hooks at the sides of the connector.
Follow the steps below to mount a removable hard disk.
*1.Flat cable with connector outfitted
Option Installation Introduction

Figure 2-1 Mounting/Removing Removable Hard Disk -1

Figure 2-2 Mounting/Removing Removable Hard Disk -2


The following is a description of the procedures to write the system correction data stored in the backup memory to the mounted removable HDD.
RestoreSysCorFile.vba from the D:\Agilent\Service folder, and then press Open button.
Figure 2-3 RestoreSysCorFile Dialog Box


When you start up the E5070B/E5071B for the first time or after executing system recovery, you need to perform the initial registration of the Windows 2000 operating system of the E5070B/E5071B.
2-4).
Figure 2-4 Windows 2000 Professional Setup wizard


box and click the Next >button (Figure 2-5). Then, the Windows 2000 operating system is restarted automatically.
Figure 2-5 Windows 2000 Professional Setup dialog box

Figure 2-6 Network Identification wizard


assumes the following user has logged on to this computer box and check that agena is in the User Name box. If not, type in agena. Finally, click the Next> button (Figure 2-7).
Figure 2-7 Network Identification Wizard dialog box (1/2)

initial registration of the E5070B/E5071B (Figure 2-8). Then, the measurement display of theE5070B/E5071B appears.
Figure 2-8 Network Identification Wizard dialog box (2/2)


When you have replaced the HDD on an E5070B/E5071B equipped with an Option 016 touch screen, you have to calibrate the touch screen. Follow the procedure described below to calibrate the touch screen.
The touch screen calibration screen (Figure 2-9) appears.
Figure 2-9 Touch Panel Calibration Screen

left, upper right, and lower right. Touch the x marks in that order with your finger.
Touching the four locations described above with your finger automatically concludes the touch screen calibration.
The RestoreSysCorFile.vba is required to write the system correction data stored in the backup memory to the orignal HDD. Execute the following steps to copy the RestoreSysCorFile.vba file.
replacing the spare HDD with the orignal HDD.
original HDD.
D:\Agilent\Service folder of the original HDD.