Agilent Technologies EMC Series Analyzers
This guide documents firmware revision A.08.xx
This manual provides documentation for the following instruments:
E7401A (9 kHz- 1.5 GHz) E7402A (9 kHz - 3.0 GHz) E7403A (9 kHz - 6.7 GHz)
E7404A (9 kHz - 13.2 GHz) E7405A (9 kHz - 26.5 GHz)

© Copyright 2001 Agilent Technologies
Printed in USA · December 2001

The information contained in this document is subject to change without notice.
Agilent Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Agilent Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
The following safety symbols are used throughout this manual. Familiarize yourself with the symbols and their meaning before operating this instrument.
The instruction documentation symbol. The product is marked with this symbol when it is necessary for the user to refer to the instructions in the documentation.
This symbol is used to mark the on position of the power line switch.
This symbol is used to mark the standby position of the power line switch.
This symbol indicates that the input power required is AC.

This Agilent Technologies instrument product is warranted against defects in material and workmanship for a period of three years from date of shipment. During the warranty period, Agilent Technologies will, at its option, either repair or replace products which prove to be defective.
For warranty service or repair, this product must be returned to a service facility designated by Agilent Technologies. Buyer shall prepay shipping charges to Agilent Technologies and Agilent Technologies shall pay shipping charges to return the product to Buyer. However, Buyer shall pay all shipping charges, duties, and taxes for products returned to Agilent Technologies from another country.
Agilent Technologies warrants that its software and firmware designated by Agilent Technologies for use with an instrument will execute its programming instructions when properly installed on that instrument. Agilent Technologies does not warrant that the operation of the instrument, or software, or firmware will be uninterrupted or error-free.

The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by Buyer, Buyer-supplied software or interfacing, unauthorized modification or misuse, operation outside of the environmental specifications for the product, or improper site preparation or maintenance.
NO OTHER WARRANTY IS EXPRESSED OR IMPLIED. AGILENT TECHNOLOGIES SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
Should Agilent have a negotiated contract with the User and should any of the contract terms conflict with these terms, the contract terms shall control.
THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND EXCLUSIVE REMEDIES. AGILENT TECHNOLOGIES SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER BASED ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY.
Documentation is updated periodically. For the latest information about Agilent Technologies EMC Analyzers, including firmware upgrades and application information, please visit the following Internet URL:
http://www.agilent.com/find/emc
Microsoft is a U.S. registered trademark of Microsoft Corporation.



Making Basic Measurements What is in This Chapter
This chapter demonstrates basic analyzer measurements with examples of typical measurements; each measurement focuses on different functions. The measurement procedures covered in this chapter are listed below.
To find descriptions of specific analyzer functions, refer to the Agilent Technologies EMC Analyzers User’s Guide.

What is in This Chapter
Making Basic Measurements

Making Basic Measurements Comparing Signals
Using the analyzer, you can easily compare frequency and amplitude differences between signals, such as radio or television signal spectra. The analyzer delta marker function lets you compare two signals when both appear on the screen at one time or when only one appears on the screen.
Measure the differences between two signals on the same display screen.
The 10 MHz reference signal appears on the display.

Figure 1-1 Placing a Marker on the 10 MHz Signal

The amplitude and frequency difference between the markers is displayed in the active function block and in the upper right corner of the screen. See Figure 1-2.
Making Basic Measurements · Comparing Signals

Making Basic Measurements Comparing Signals
Figure 1-2 Using the Marker Delta Function

Measure the frequency and amplitude difference between two signals that do not appear on the screen at one time. (This technique is useful for harmonic distortion tests when narrow span and narrow bandwidth are necessary to measure the low level harmonics.)
The 10 MHz reference signal appears on the display.

The annotation in the upper right corner of the screen indicates the amplitude and frequency difference between the two markers.
Figure 1-3 Frequency and Amplitude Difference Between Signals

Making Basic Measurements · Comparing Signals

Making Basic Measurements Resolving Signals of Equal Amplitude
Two equal-amplitude input signals that are close in frequency can appear as a single signal trace on the analyzer display. Responding to a single-frequency signal, a swept-tuned analyzer traces out the shape of the selected internal IF (intermediate frequency) filter. As you change the filter bandwidth, you change the width of the displayed response. If a wide filter is used and two equal-amplitude input signals are close enough in frequency, then the two signals will appear as one signal. If a narrow enough filter is used, the two input signals can be discriminated and will appear as separate peaks. Thus, signal resolution is determined by the IF filters inside the analyzer.
The bandwidth of the IF filter tells us how close together equal amplitude signals can be and still be distinguished from each other. The resolution bandwidth function selects an IF filter setting for a measurement. Typically, resolution bandwidth is defined as the 3 dB bandwidth of the filter. However, resolution bandwidth may also be defined as the 6 dB or impulse bandwidth of the filter.
Generally, to resolve two signals of equal amplitude, the resolution bandwidth must be less than or equal to the frequency separation of the two signals. If the bandwidth is equal to the separation and the video bandwidth is less than the resolution bandwidth, a dip of approximately 3 dB is seen between the peaks of the two equal signals, and it is clear that more than one signal is present. See Figure 1-7.
In order to keep the analyzer measurement calibrated, sweep time is automatically set to a value that is inversely proportional to the square of the resolution bandwidth (1/BW2 for resolution bandwidths ≥ 1 kHz). So, if the resolution bandwidth is reduced by a factor of 10, the sweep time is increased by a factor of 100 when sweep time and bandwidth settings are coupled. Sweep time is also a function of the type of detection selected (peak detection is faster than sample or average detection). For the shortest measurement times, use the widest resolution bandwidth that still permits discrimination of all desired signals. Sweeptime is also a function of which Detector is in use, Peak detector sweeps more quickly than Sample or Average detector. The analyzer allows you to select from 10 Hz (or 1 Hz with Option 1D5) to 3 MHz resolution bandwidths in a 1, 3, 10 sequence and select a 5 MHz resolution bandwidth. In addition you can select the three CISPR bandwidths (200 Hz, 9 kHz, and 120 kHz) for maximum measurement flexibility.

Making Basic Measurements Resolving Signals of Equal Amplitude
Resolve two signals of equal amplitude with a frequency separation of 100 kHz.
Figure 1-4 Setup for Obtaining Two Signals

a. Press Preset, Factory Preset (if present).
b. Set the Y-Axis Units to dBm by pressing AMPLITUDE, More,
Y-Axis Units, dBm.
c. Set the center frequency to 300 MHz by pressing FREQUENCY,
Center Freq, 300, MHz.
d. Set the span to 2 MHz by pressing SPAN, Span, 2, MHz.
e. Set the resolution bandwidth to 300 kHz by pressing BW/Avg,
Res BW, 300, kHz.
A single signal peak is visible. See Figure 1-5

Making Basic Measurements Resolving Signals of Equal Amplitude
Figure 1-5 Unresolved Signals of Equal Amplitude

4. Since the resolution bandwidth must be less than or equal to the frequency separation of the two signals, a resolution bandwidth of 100 kHz must be used. Change the resolution bandwidth to 100 kHz by pressing BW/Avg, Res BW, 100, kHz. The peak of the signal has become flattened indicating that two signals may be present as shown in Figure 1-6. Use the knob or step keys to further reduce the resolution bandwidth and better resolve the signals.
Figure 1-6 Resolving Signals of Equal Amplitude Before Reducing the Video Bandwidth


Making Basic Measurements Resolving Signals of Equal Amplitude
Figure 1-7 Resolving Signals of Equal Amplitude After Reducing the Video Bandwidth

As the resolution bandwidth is decreased, resolution of the individual signals is improved and the sweep time is increased. For fastest measurement times, use the widest possible resolution bandwidth. Under factory preset conditions, the resolution bandwidth is “coupled” (or linked) to the center frequency.
Since the resolution bandwidth has been changed from the coupled value, a # mark appears next to Res BW in the lower-left corner of the screen, indicating that the resolution bandwidth is uncoupled. (For more information on coupling, refer to the Auto Couple key description in the Agilent Technologies EMC Analyzers User’s Guide.)

Making Basic Measurements Resolving Small Signals Hidden by Large Signals
When dealing with the resolution of signals that are close together and not equal in amplitude, you must consider the shape of the IF filter of the analyzer, as well as its 3 dB bandwidth. (See “Resolving Signals of Equal Amplitude” on page 14 for more information.) The shape of a filter is defined by the selectivity, which is the ratio of the 60 dB bandwidth to the 3 dB bandwidth. (Generally, the IF filters in this analyzer have shape factors of 15:1 or less for resolution bandwidths ≥1 kHz and 5:1 or less for resolution bandwidths ≤ 300 Hz). If a small signal is too close to a larger signal, the smaller signal can be hidden by the skirt of the larger signal. To view the smaller signal, you must select a resolution bandwidth such that the separation between the two signals (a) is greater than half the filter width of the larger signal (k) measured at the amplitude level of the smaller signal. See Figure 1-8.
Figure 1-8 Resolution Bandwidth Requirements for Resolving Small Signals


Making Basic Measurements Resolving Small Signals Hidden by Large Signals
Resolve two input signals with a frequency separation of 155 kHz and an amplitude separation of 60 dB.
Figure 1-9 Setup for Obtaining Two Signals
a. Press Preset, Factory Preset (if present).
b. Set the Y-Axis Units to dBm by pressing AMPLITUDE, More,
Y-Axis Units, dBm.
c. Set the center frequency to 300 MHz by pressing FREQUENCY,
Center Freq, 300, MHz.
d. Set the span to 2 MHz by pressing SPAN, Span, 2, MHz.
e. Set the resolution bandwidth to spectrum analyzer coupling by
pressing BW/Avg, Resolution BW (SA).

Making Basic Measurements Resolving Small Signals Hidden by Large Signals
If a 10 kHz filter with a typical shape factor of 15:1 is used, the filter will have a bandwidth of 150 kHz at the 60 dB point. The half-bandwidth (75 kHz) is narrower than the frequency separation, so the input signals will be resolved. See Figure 1-10.
Figure 1-10 Signal Resolution with a 10 kHz Resolution Bandwidth

Figure 1-11 Signal Resolution with a 10 kHz Resolution Bandwidth


Making Basic Measurements Resolving Small Signals Hidden by Large Signals
When a 30 kHz filter is used, the 60 dB bandwidth could be as wide as 450 kHz. Since the half-bandwidth (225 kHz) is wider than the frequency separation, the signals most likely will not be resolved. See Figure 1-12. (In this example, we used the 60 dB bandwidth value. To determine resolution capability for intermediate values of amplitude level differences, assume the filter skirts between the 3 dB and 60 dB points are approximately straight.)
Figure 1-12 Signal Resolution with a 30 kHz Resolution Bandwidth


Making Basic Measurements Making Better Frequency Measurements
A built-in frequency counter increases the resolution and accuracy of the frequency readout. When using this function, if the ratio of the resolution bandwidth to the span is too small (less than 0.002), the Marker Count: Widen Res BW message appears on the display. It indicates that the resolution bandwidth is too narrow.
Increase the resolution and accuracy of the frequency readout on the signal of interest.

Making Basic Measurements Making Better Frequency Measurements
components. The marker must be >26 dB above the noise.
Figure 1-13 Using Marker Counter


Making Basic Measurements Decreasing the Frequency Span Around the Signal
Using the analyzer signal track function, you can quickly decrease the span while keeping the signal at center frequency. This is a fast way to take a closer look at the area around the signal to identify signals that would otherwise not be resolved.
Examine a signal in a 200 kHz span.

Making Basic Measurements Decreasing the Frequency Span Around the Signal
Figure 1-14 Detected Signal

Because the signal track function automatically maintains the signal at the center of the screen, you can reduce the span quickly for a closer look. If the signal drifts off of the screen as you decrease the span, use a wider frequency span. (You can also use Span Zoom, in the SPAN menu, as a quick way to perform the Peak Search, FREQUENCY, Signal Track, SPAN key sequence.)
Figure 1-15 Signal with Signal Tracking On


Making Basic Measurements Decreasing the Frequency Span Around the Signal
If the span change is large enough, span will decrease in steps as automatic zoom is completed. See Figure 1-16. You can also use the front-panel knob or step keys to decrease the span and resolution bandwidth values.
Figure 1-16 After Zooming In on the Signal


The signal track function is useful for tracking drifting signals that drift relatively slowly. To place a marker on the signal you wish to track, use Peak Search. Pressing FREQUENCY, Signal Track (On) will bring that signal to the center frequency of the graticule and adjust the center frequency every sweep to bring the selected signal back to the center. A quick way to perform the Peak Search, FREQUENCY, Signal Track, SPAN key sequence is to use the Span Zoom key in the SPAN menu.
Use the signal track function to keep a drifting signal at the center of the display and monitor its change.
This example requires a signal generator. The frequency of the signal generator will be changed while you view the signal on the display of the analyzer.
a. Press Preset, Factory Preset (if present).
b. Set the Y-Axis Units to dBm by pressing AMPLITUDE, More,
Y-Axis Units, dBm.
c. Set the resolution bandwidth to the spectrum analyzer coupling
by pressing BW/Avg, Resolution BW (SA). See Figure 1-17.
d. Set the center frequency to 300 MHz by pressing FREQUENCY,
Center Freq, 300, MHz.
Making Basic Measurements · Tracking Drifting Signals

Making Basic Measurements Tracking Drifting Signals
Figure 1-17 Signal With Default Span

Figure 1-18 Signal With 10 MHz Span


Figure 1-19 Signal With 500 kHz Span

Figure 1-20 Using Span Zoom to Track a Drifting Signal

Making Basic Measurements · Tracking Drifting Signals

Making Basic Measurements Tracking Drifting Signals
a. Press Marker, Delta.
b. Tune the frequency of the signal generator. The marker readout
indicates the change in frequency and amplitude as the signal drifts. See Figure 1-21.
Figure 1-21 Using Signal Tracking to Track a Drifting Signal

The analyzer can measure the short- and long-term stability of a source. The maximum amplitude level and the frequency drift of an input signal trace can be displayed and held by using the maximum-hold function. You can also use the maximum hold function if you want to determine how much of the frequency spectrum a signal occupies.
a. Press Preset, Factory Preset (if present).
b. Set the Y-Axis Units to dBm by pressing AMPLITUDE, More,
Y-Axis Units, dBm.
c. Set the resolution bandwidth to spectrum analyzer coupling by
pressing BW/Avg, Resolution BW (SA).

d. Set the center frequency to 300 MHz by pressing FREQUENCY,
Center Freq, 300, MHz. See Figure 1-22.
Figure 1-22 Signal With Default Span

Figure 1-23 Signal With 10 MHz Span

Making Basic Measurements · Tracking Drifting Signals

Making Basic Measurements Tracking Drifting Signals
Figure 1-24 Signal With 500 KHz Span

Annotation on the left side of the screen indicates the trace mode. For example, M1 S2 S3 indicates trace 1 is in maximum-hold mode, trace 2 and trace 3 are in store-blank mode.

Figure 1-25 Viewing a Drifting Signal With Max Hold and Clear Write

Making Basic Measurements · Tracking Drifting Signals

Making Basic Measurements Measuring Low Level Signals
The ability of the analyzer to measure low level signals is limited by the noise generated inside the analyzer. A signal may be masked by the noise floor so that it is not visible. This sensitivity to low level signals is affected by the measurement setup.
The analyzer input attenuator and bandwidth settings affect the sensitivity by changing the signal-to-noise ratio. The attenuator affects the level of a signal passing through the instrument, whereas the bandwidth affects the level of internal noise without affecting the signal. In the first two examples in this section, the attenuator and bandwidth settings are adjusted to view low level signals.
If, after adjusting the attenuation and resolution bandwidth, a signal is still near the noise, visibility can be improved by using the video bandwidth and video averaging functions, as demonstrated in the third and fourth examples.
If a signal is very close to the noise floor, reducing input attenuation brings the signal out of the noise. Reducing the attenuation to 0 dB maximizes signal power in the analyzer.

Making Basic Measurements Measuring Low Level Signals
Figure 1-26 Low-Level Signal

A # mark appears next to the Atten annotation at the top of the display, indicating the attenuation is no longer coupled to other analyzer settings.
Figure 1-27 Using 20 dB Attenuation


Making Basic Measurements Measuring Low Level Signals
Figure 1-28 Using 0 dB Attenuation

The resolution bandwidth can be decreased to view low level signals.

Making Basic Measurements Measuring Low Level Signals
A # mark appears next to the Res BW annotation at the lower left corner of the screen, indicating that the resolution bandwidth is uncoupled. As the resolution bandwidth is reduced, the sweep time is increased to maintain calibrated data.
Figure 1-29 Decreasing Resolution Bandwidth

Narrowing the video filter can be useful for noise measurements and observation of low level signals close to the noise floor. The video filter is a post-detection low-pass filter that smooths the displayed trace. When signal responses near the noise level of the analyzer are visually masked by the noise, the video filter can be narrowed to smooth this noise and improve the visibility of the signal. (Reducing video bandwidths requires slower sweep times to keep the analyzer calibrated.)
Using the video bandwidth function, measure the amplitude of a low level signal.

Making Basic Measurements Measuring Low Level Signals
Figure 1-30 30 kHz Video Bandwidth

A # mark appears next to the VBW annotation at the bottom of the screen, indicating that the video bandwidth is not coupled to the resolution bandwidth. See Figure 1-31. As the video bandwidth is reduced, the sweep time is increased to maintain calibrated data.
Instrument preset conditions couple the video bandwidth to the resolution bandwidth. If the bandwidths are uncoupled when video bandwidth is the active function, pressing Video BW (so that Auto is underlined) recouples the bandwidths.

Making Basic Measurements Measuring Low Level Signals
Figure 1-31 Decreasing Video Bandwidth

If a signal level is very close to the noise floor, video averaging is another way to make the signal more visible.
NOTE The time required to construct a full trace that is averaged to the desired degree is approximately the same when using either the video bandwidth or the video averaging technique. The video bandwidth technique completes the averaging as a slow sweep is taken, whereas the video averaging technique takes many sweeps to complete the average. Characteristics of the signal being measured, such as drift and duty cycle, determine which technique is appropriate.
Video averaging is a digital process in which each trace point is averaged with the previous trace-point average. Selecting Average Type, Video Avg and Average (On) changes the detection mode from peak to sample. The result is a sudden drop in the displayed noise level. The sample mode displays the instantaneous value of the signal at the end of the time or frequency interval represented by each display point, rather than the value of the peak during the interval. Sample mode is not used to measure signal amplitudes accurately because it may not find the true peak of the signal.
Video averaging clarifies low-level signals in wide bandwidths by averaging the signal and the noise. As the analyzer takes sweeps, you can watch video averaging smooth the trace.

Making Basic Measurements Measuring Low Level Signals
Figure 1-32 Without Video Averaging


Making Basic Measurements Measuring Low Level Signals
During averaging, the current sample number appears at the left side of the graticule. The number of samples equals the number of sweeps in the averaging routine. Changes in active function settings, such as the center frequency or reference level, will restart the sampling. The sampling will also restart if video averaging is turned off and then on again. To see the sample number increment, turn video averaging off and on again by pressing Average (Off), Average (On).
Figure 1-33 Using the Video Averaging Function


Making Basic Measurements Identifying Distortion Products
High level input signals may cause analyzer distortion products that could mask the real distortion measured on the input signal. Using trace 2 and the RF attenuator, you can determine which signals, if any, are internally generated distortion products.
Using a signal from a signal generator, determine whether the harmonic distortion products are generated by the analyzer.
The signal produces harmonic distortion products in the analyzer input mixer as shown in Figure 1-34.

Making Basic Measurements Identifying Distortion Products
Figure 1-34 Harmonic Distortion

Figure 1-35 Harmonic Distortion with 0 dB Attenuation


Making Basic Measurements Identifying Distortion Products
a. Press Trace/View, Trace (2), then Clear Write.
b. Allow the trace to update (two sweeps) and press Trace/View, View,
Marker, Delta. The analyzer display shows the stored data in trace 2 and the measured data in trace 1.
Notice the ∆Mkr1 amplitude reading. This is the difference in the distortion product amplitude readings between 0 dB and 10 dB input attenuation settings. If the ∆Mkr1 amplitude absolute value is approximately ≥1 dB for an input attenuator change, the distortion is being generated, at least in part, by the analyzer. In this case more input attenuation is necessary.
Figure 1-36 RF Attenuation of 10 dB

Change the attenuation to 15 dB by pressing Attenuation, 15, dB.
If the ∆Mkr1 amplitude absolute value is approximately ≥1 dB as seen in Figure 1-36, then more input attenuation is required; some of the measured distortion is internally generated. If there is no change in the signal level, the distortion is not generated internally. For example, the signal that is causing the distortion shown in Figure 1-37 is not high enough in amplitude to cause internal distortion in the analyzer so any distortion that is displayed is present on the input signal.

Making Basic Measurements Identifying Distortion Products
Figure 1-37 No Harmonic Distortion

Two-tone, third-order intermodulation distortion is a common test in communication systems. When two signals are present in a non-linear system, they can interact and create third-order intermodulation distortion products that are located close to the original signals. These distortion products are generated by system components such as amplifiers and mixers.
Test a device for third-order intermodulation. This example uses two sources, one set to 300 MHz and the other to approximately 301 MHz. (Other source frequencies may be substituted, but try to maintain a frequency separation of approximately 1 MHz.)
1. Connect the equipment as shown in Figure 1-38. This combination of signal generators, low pass filters, and directional coupler (used as a combiner) results in a two-tone source with very low intermodulation distortion. Although the distortion from this setup may be better than the specified performance of the analyzer, it is useful for determining the TOI performance of the source/analyzer combination. After the performance of the source/analyzer combination has been verified, the device-under-test (DUT) (for example, an amplifier) would be inserted between the directional coupler output and the analyzer input and another measurement would be made.

Making Basic Measurements Identifying Distortion Products
Figure 1-38 Third-Order Intermodulation Equipment Setup

If necessary, use the front-panel knob to center the two test signals on the display.

Making Basic Measurements Identifying Distortion Products
The analyzer automatically sets the attenuation so that a signal at the reference level will be a maximum of −30 dBm at the input mixer.
Figure 1-39 Measuring the Distortion Product


Making Basic Measurements Identifying Distortion Products
Figure 1-40 Measuring the Distortion Product


The signal-to-noise measurement procedure below may be adapted to measure any signal in a system if the signal (carrier) is a discrete tone. If the signal in your system is modulated, it will be necessary to modify the procedure to correctly measure the modulated signal level
In this example the 50 MHz amplitude reference signal is used as the fundamental source. The amplitude reference signal is assumed to be the signal of interest and the internal noise of the analyzer is measured as the system noise. To do this, you will need to set the input attenuator such that both the signal and the noise are well within the calibrated region of the display.
Perform the steps below to measure the signal-to-noise.
Making Basic Measurements · Measuring Signal-to-Noise

Making Basic Measurements Measuring Signal-to-Noise
Figure 1-41 Measuring the Signal-to-Noise

Read the signal-to-noise in dB/Hz, that is with the noise value determined for a 1 Hz noise bandwidth. If you wish the noise value for a different bandwidth, decrease the ratio by . For example, if the analyzer reading is −70 dB/Hz but you have a channel bandwidth of 30 kHz:
10 log × BW ( )
S/N 70 dB/Hz – 10 30 kHz ( ) log × + 25.23 dB – 30 kHz ( ) ⁄ = =

Making Basic Measurements Making Noise Measurements
There are a variety of ways to measure noise power. The first decision you must make is whether you want to measure noise power at a specific frequency or the total power over a specified frequency range, for example over a channel bandwidth.
Using the marker function, Marker Noise, is a simple method to make a measurement at a single frequency. In this example, attention must be made to the potential errors due to discrete signal (spectral components). This measurement will be made near the 50 MHz amplitude reference signal to illustrate the use of Marker Noise.

Making Basic Measurements Making Noise Measurements
Figure 1-42 Setting the Attenuation

Note that the display detection automatically changed to “Avg” which can be manually set by pressing Det/Demod, Average (Video/RMS). The marker is floating between the maximum and the minimum of the noise. For firmware revisions earlier than A.08.00, the detection type when using Marker Noise changed to sample. If you wish to use sample detection, press Det/Demod, Detector, Sample and verify that “Average Type” is set to “Video average” by pressing BW/Avg, Average Type, Video Avg. This is not recommended as it is slower and does not increase accuracy.
The marker readout is in dBm(Hz) or dBm per unit bandwidth. See Figure 1-43. For noise power in a different bandwidth, add
10 log × BW ( )
. For example, for noise power in a 1 kHz bandwidth, add
10 log 1000 ( ) ×
or 30 dB to the noise marker value.

Making Basic Measurements Making Noise Measurements
Figure 1-43 Activating the Noise Marker

Figure 1-44 Noise Marker at 50 MHz


Making Basic Measurements Making Noise Measurements
Figure 1-45 Increased Resolution Bandwidth


Making Basic Measurements Making Noise Measurements
Figure 1-46 Noise Marker in Signal Skirt

Figure 1-47 Noise Marker with Zero Span


Making Basic Measurements Making Noise Measurements
The Normal marker can also be used to make a single frequency measurement as described in the previous example, again using video filtering or averaging to obtain a reasonably stable measurement. While video averaging automatically selects the sample display detection mode, video filtering does not. With sufficient filtering that results in a smooth trace, there is no difference between the sample and peak modes because the filtering takes place before the signal is digitized.
Be sure to account for the fact that the averaged noise is displayed approximately 2 dB too low for a noise bandwidth equal to the resolution bandwidth. Therefore, you must add 2 dB to the marker reading. For example, if the marker indicates –100 dBm, the actual noise level is –98 dBm.
You may use adjustable markers to set the frequency span over which power is measured. The markers allow you to easily and conveniently select any arbitrary portion of the displayed signal for measurement. However, while the analyzer does select the average (video/rms) display detection mode, you must set all of the other parameters.
The resolution bandwidth should be about 1 to 3% of the measurement (marker) span, 40 kHz in this example. The 1 kHz resolution bandwidth that the analyzer has chosen is fine. The video bandwidth should be ten times wider.

Making Basic Measurements Making Noise Measurements
Figure 1-48 Viewing Power Between Markers


Making Basic Measurements Making Noise Measurements
Figure 1-49 Measuring the Power in the Span


Making Basic Measurements Demodulating AM Signals (Using the Analyzer As a Fixed Tuned Receiver)
The zero span mode can be used to recover amplitude modulation on a carrier signal. The analyzer operates as a fixed-tuned receiver in zero span to provide time domain measurements.
Center frequency in the swept-tuned mode becomes the tuned frequency in zero span. The horizontal axis of the screen becomes calibrated in time only, rather than both frequency and time. Markers display amplitude and time values.
The following functions establish a clear display of the waveform:
Each of the coupled function values remains at its current value when zero span is activated. Video bandwidth is coupled to resolution bandwidth. Sweep time is not coupled to any other function.
To obtain an AM signal, you can either connect a source to the analyzer input and set the source for amplitude modulation, or connect an antenna to the analyzer input and tune to a commercial AM broadcast station.
View the modulation waveform of an AM signal in the time domain.
a. RF Frequency 300 MHz

Making Basic Measurements Demodulating AM Signals (Using the Analyzer As a Fixed Tuned Receiver)
b. RF Output Power –10 dBm
c. AM On
d. AM Rate 1 kHz
e. AM Depth 80%
a. Press Preset, Factory Preset (if present).
b. Set the center frequency to 300 MHz by pressing FREQUENCY,
Center Freq, 300, MHz.
c. Set the span to 500 kHz by pressing SPAN, Span, 500, kHz.
d. Set the resolution bandwidth to 30 kHz by pressing BW/Avg,
Resolution BW, 30, kHz.
e. Set the Y-Axis Units to dBm by pressing AMPLITUDE, More,
Y-Axis Units, dBm.
f. Change the analyzer sweep to 20 msec by pressing Sweep, Sweep Time, 20, ms. See Figure 1-50.
Figure 1-50 Viewing an AM Signal


Making Basic Measurements Demodulating AM Signals (Using the Analyzer As a Fixed Tuned Receiver)
If you are viewing an off-the-air signal you will not be able to stabilize the waveform.
Figure 1-51 Measuring Modulation In Zero Span


Making Basic Measurements Demodulating AM Signals (Using the Analyzer As a Fixed Tuned Receiver)
Figure 1-52 Measuring Modulation In Zero Span

Figure 1-53 Measuring Modulation In Zero Span

a. Press Marker and center the marker on a peak using Peak Search
or the front-panel knob.
b. Press Marker, Delta and center the marker on the next peak using
the front-panel knob or use Peak Search and Next Pk Right (or Next Pk Left). See Figure 1-54.

Making Basic Measurements Demodulating AM Signals (Using the Analyzer As a Fixed Tuned Receiver)
Figure 1-54 Measuring Time Parameters

a. Set trigger to free run by pressing Trig, Free Run.
b. Set the sweep time to 5 seconds by pressing Sweep, Sweep Time, 5,
s.
c. Set the video filter to 30 Hz by pressing BW/Avg, Video BW, 30, Hz.
d. Change the reference level to position the trace at midscreen by
pressing AMPLITUDE, Ref Level, and adjacent the reference level using the front-panel knob.
e. Reset the video filter to a high value. For example, press BW/Avg,
Video BW, 100, kHz.
f. Set the sweep time to 5 milliseconds by pressing Sweep, Sweep Time, 5, ms.
The center horizontal line of the graticule now represents 0% AM; the top and bottom lines, 100% AM. See Figure 1-55.

Making Basic Measurements Demodulating AM Signals (Using the Analyzer As a Fixed Tuned Receiver)
Figure 1-55 Continuous Demodulation of an AM Signal


As with amplitude modulation (see page 59) you can utilize zero span to demodulate an FM signal. However, unlike the AM case, you cannot simply tune to the carrier frequency and widen the resolution bandwidth. The reason is that the envelope detector in the analyzer responds only to amplitude variations, and there is no change in amplitude if the frequency changes of the FM signal are limited to the flat part of the resolution bandwidth.
On the other hand, if you tune the analyzer slightly away from the carrier, you can utilize slope detection to demodulate the signal by performing the following steps.
The demodulated signal is now displayed; the frequency changes have been translated into amplitude changes., see Figure 1-58. To listen to the signal, turn on AM demodulation and the speaker.
In this example you will demodulate a broadcast FM signal that has a specified 75 kHz peak deviation.
Determine the correct resolution bandwidth. With a peak deviation of 75 kHz, your signal has a peak-to-peak excursion of 150 kHz. So we must find a resolution bandwidth filter with a skirt that is reasonably linear over that frequency range.
Making Basic Measurements · Demodulating FM Signals

Making Basic Measurements Demodulating FM Signals
Figure 1-56 Establishing the Offset Point


Figure 1-57 Determining the Offset

Demodulate the FM Signal
Making Basic Measurements · Demodulating FM Signals

Making Basic Measurements Demodulating FM Signals
Figure 1-58 Demodulating a Broadcast Signal



Making Complex Measurements What’s in This Chapter
This chapter provides information for making complex measurements. The procedures covered in this chapter are listed below.
To find descriptions of specific analyzer functions refer to the Agilent Technologies EMC Series Analyzers User’s Guide.
Test Equipment Specifications Recommended Model
Type-N (m) to BNC (f) (2) 1250-0780
(3) BNC, 122-cm (48-in) 10503A
Adapters · Cables · Miscellaneous

Making Complex Measurements Making Stimulus Response Measurements
Stimulus response measurements require a source to stimulate a device under test (DUT), a receiver to analyze the frequency response characteristics of the DUT, and, for return loss measurements, a directional coupler or bridge. Characterization of a DUT can be made in terms of its transmission or reflection parameters. Examples of transmission measurements include flatness and rejection. Return loss is an example of a reflection measurement.
A spectrum analyzer combined with a tracking generator forms a stimulus response measurement system. With the tracking generator as the swept source and the analyzer as the receiver, operation is the same as a single channel scalar network analyzer. The tracking generator output frequency must be made to precisely track the analyzer input frequency for good narrow band operation. A narrow band system has a wide dynamic measurement range. This wide dynamic range will be illustrated in the following example.
There are three basic steps in performing a stimulus response measurement, whether it is a transmission or a reflection measurement. The steps are to set all the analyzer settings, normalize, and measure.
The procedure below describes how to use a built in tracking generator system to measure the rejection of a band pass filter, a type of transmission measurement. Illustrated in this example are functions in the tracking generator menu such as adjusting the tracking generator output power. Normalization functions located in the trace menu are also used. Making a reflection measurement is similar and is covered in “Making a Reflection Calibration Measurement” on page 84.

Making Complex Measurements Making Stimulus Response Measurements
Figure 2-1 Transmission Measurement Test Setup

a. Set the span to 100 MHz by pressing SPAN, Span, 100, MHz.
b. Set the center frequency to 200 MHz by pressing FREQUENCY,
Center Freq, 200, MHz.
NOTE To reduce ripples caused by source return loss, use 10 dB (E7401A) or 8 dB (all other models) or greater tracking generator output attenuation. Tracking generator output attenuation is normally a function of the source power selected. However, the output attenuation may be controlled in the Source menu. Refer to specifications and characteristics in your specifications guide for more information on the relationship between source power and source attenuation.

Making Complex Measurements Making Stimulus Response Measurements
Figure 2-2 Tracking Generator Output Power Activated

NOTE In the stimulus response mode, the Q of the DUT can determine the fastest rate at which the analyzer can be swept. (Q is the quality factor, which is the center frequency of the DUT divided by the bandwidth of the DUT.) To determine whether the analyzer is sweeping too fast, slow the sweep and note whether there is a frequency or amplitude shift of the trace. Continue to slow the sweep until there is no longer a frequency or amplitude shift.

Making Complex Measurements Making Stimulus Response Measurements
Figure 2-3 Decrease the Resolution Bandwidth to Improve Sensitivity

10.You might notice a decrease in the displayed amplitude as the resolution bandwidth is decreased, (if the analyzer is an E7402A, E7403A, E7404A, or E7405A). This indicates the need for performing a tracking peak. Press Source, Tracking Peak. The amplitude should return to that which was displayed prior to the decrease in resolution bandwidth.
11.To make a transmission measurement accurately, the frequency response of the test system must be known. Normalization is used to eliminate this error from the measurement. To measure the frequency response of the test system, connect the cable (but not the DUT) from the tracking generator output to the analyzer input. Press Trace/View, More, Normalize, Store Ref (1→3), Normalize (On). The frequency response of the test system is automatically stored in trace 3 and a normalization is performed. This means that the active displayed trace is now the ratio of the input data to the data stored in trace 3. (The reference trace is Trace 3 with firmware revision A.04.00 and later)
When normalization is on, trace math is being performed on the active trace. The trace math performed is (trace 1 − trace 3 + the normalized reference position), with the result placed into trace 1. Remember that trace 1 contains the measurement trace, trace 3 contains the stored reference trace of the system frequency response, and normalized reference position is indicated by arrowheads at the edges of the graticule.

Making Complex Measurements Making Stimulus Response Measurements
12.Reconnect the DUT to the analyzer. Note that the units of the reference level have changed to dB, indicating that this is now a relative measurement.
Press Trace/View, More, Normalize, Norm Ref Posn to change the normalized reference position. Arrowheads at the left and right edges of the graticule mark the normalized reference position, or the position where 0 dB insertion loss (transmission measurements) or 0 dB return loss (reflection measurements) will normally reside. Using the knob results in a change in the position of the normalized trace, within the range of the graticule.
13.To measure the rejection of the filter 45 MHz above the center of the bandpass, press Marker, 200, MHz (to ensure that the marker is in the center of the signal), and then press Delta, 45, MHz. The marker readout displays the rejection of the filter at 45 MHz above the center of the bandpass. See Figure 2-4.
NOTE Because the default trace is comprised of 401 discrete points, the indicated marker frequency may differ slightly from the frequency that you entered. Due to the horizontal resolution of the trace, the marker frequency value will be rounded to within 0.25% of the span of the value entered. If the analyzer is an ESA-E series with firmware revision A.04.00 or later, the number of sweep points may be set to any value between 101 and 8192.
Figure 2-4 Measure the Rejection Range


Making Complex Measurements Making Stimulus Response Measurements
When using the tracking generator, the message TG unleveled may appear. The TG unleveled message indicates that the tracking generator source power (Source, Amplitude) could not be maintained at the selected level during some portion of the sweep. If the unleveled condition exists at the beginning of the sweep, the message will be displayed immediately. If the unleveled condition occurs after the sweep begins, the message will be displayed after the sweep is completed. A momentary unleveled condition may not be detected when the sweep time is short. The message will be cleared after a sweep is completed with no unleveled conditions.
The unleveled condition may be caused by any of the following:
It is often necessary to measure device bandwidth, such as when testing a bandpass filter. There is a key in the Peak Search menu that will perform this function. The device signal being measured must be displayed before activating the measurement. The span must include the full response.
Activate the measurement by toggling the N dB Points key to On. The analyzer places arrow markers at the −3 dB points on either side of the response and reads the bandwidth. For other bandwidth responses enter the number of dB down desired, from −1 dB to −80 dB.
No other signal can appear on the display within N dB of the highest signal. The measured signal cannot have more than one peak that is greater than or equal to N dB. A signal must have a peak greater than the currently defined peak excursion to be identified. The default value for the peak excursion is 6 dB.

Making Complex Measurements Making Stimulus Response Measurements
Measurements are made continuously, updating at the end of each sweep. This allows you to make adjustments and see changes as they happen. The single sweep mode can also be used, providing time to study or record the data.
The N dB bandwidth measurement error is typically ±1% of the span.
Example:
Measure the 3 dB bandwidth of a 200 MHz bandpass filter.
Figure 2-5 Transmission Measurement Test Setup

Making Complex Measurements Making Stimulus Response Measurements
NOTE To reduce ripples caused by source return loss, use 10 dB (E7401A) or 8 dB (all other models) or greater tracking generator output attenuation. Tracking generator output attenuation is normally a function of the source power selected. However, the output attenuation may be controlled in the Source menu. Refer to specifications and characteristics in your specifications guide for more information on the relationship between source power and source attenuation.
NOTE In the stimulus response mode, the Q of the DUT can determine the fastest rate at which the analyzer can be swept. (Q is the quality factor, which is the center frequency of the DUT divided by the bandwidth of the DUT.) To determine whether the analyzer is sweeping too fast, slow the sweep and note whether there is a frequency or amplitude shift of the trace. Continue to slow the sweep until there is no longer a frequency or amplitude shift.
10.Read the measurement results displayed on the screen.
Figure 2-6 N dB Bandwidth Measurement at –3 dB


Making Complex Measurements Making Stimulus Response Measurements
11.The knob or the data entry keys can be used to change the N dB value from −3 dB to −60 dB to measure the 60 dB bandwidth of the filter. See Figure 2-7.
Figure 2-7 N dB Bandwidth Measurement at –60 dB

12.Press N dB Points (Off) to turn the measurement off.
When measuring filter characteristics, it is useful to look at the stimulus response over a wide frequency range. Setting the analyzer x-axis (frequency) to display logarithmically provides this function.
Example:
Measure the stop band attenuation of a 10 MHz low pass filter.

Making Complex Measurements Making Stimulus Response Measurements
Figure 2-8 Transmission Measurement Test Setup
NOTE To reduce ripples caused by source return loss, use 10 dB (E7401A) or 8 dB (all other models) or greater tracking generator output attenuation. Tracking generator output attenuation is normally a function of the source power selected. However, the output attenuation may be controlled in the Source menu. Refer to specifications and characteristics in your specifications guide for more information on the relationship between source power and source attenuation.

Making Complex Measurements Making Stimulus Response Measurements
Figure 2-9 Tracking Generator Output Power Activated in Log Sweep

10.Put the sweep time of the analyzer into stimulus response auto coupled mode by pressing Sweep, then Swp Coupling (SR). See Figure 2-9. Auto coupled sweep times are usually much faster for stimulus response measurements than they are for spectrum analyzer (SA) measurements. Adjust the reference level if necessary to place the signal on screen.
11.To make a transmission measurement accurately, the frequency response of the test system must be known. Normalization is used to eliminate this error from the measurement. To measure the frequency response of the test system, connect the cable (but not the DUT) from the tracking generator output to the analyzer input. Press Trace/View, More, Normalize, Store Ref (1→3), Normalize (On).
This will activate the trace 1 minus trace 3 function and display the results in trace 1. The normalized trace or flat line represents 0 dB return loss. Normalization occurs each sweep.
12.Reconnect the DUT to the analyzer. Note that the units of the reference level have changed to dB, indicating that this is now a relative measurement. Refer to Figure 2-10.
13.Press Trace/View, More, Normalize, Norm Ref Posn to change the normalized reference position. Arrowheads at the left and right edges of the graticule mark the normalized reference position. Using the knob results in a change in the position of the normalized trace, within the range of the graticule. Refer to Figure 2-10.

Making Complex Measurements Making Stimulus Response Measurements
Figure 2-10 Normalized Trace After Reconnecting DUT

14.Press Marker, Delta Pair (Ref), 10, MHz to place the reference marker at the specified cutoff frequency.
15.Press Delta Pair (∆), 20, MHz to place the second marker at the 20 MHz point. In this example, the attenuation over this frequency range is 63.32 dB/octave (one octave above the cutoff frequency).
Figure 2-11 Determining Low Pass Filter Rolloff

16.Use the knob to place the marker at the highest peak in the stop band to determine the minimum stop band attenuation. In this example, the peak occurs at 708.76 MHz. The attenuation is 54.92 dB.

Making Complex Measurements Making Stimulus Response Measurements
Figure 2-12 Minimum Stop Band Attenuation


Making Complex Measurements Making a Reflection Calibration Measurement
The calibration standard for reflection measurements is usually a short circuit connected at the reference plane (the point at which the device under test (DUT) will be connected.) See Figure 2-13. A short circuit has a reflection coefficient of 1 (0 dB return loss). It reflects all incident power and provides a convenient 0 dB reference.
Figure 2-13 Reflection Measurement Short Calibration Test Setup

Measure the return loss of a filter. The following procedure makes a reflection measurement using a coupler or directional bridge. This example uses a 200 MHz bandpass filter as the DUT.

Making Complex Measurements Making a Reflection Calibration Measurement
10.Replace the DUT with a short circuit.
11.Normalize the trace by pressing Trace/View, More, Normalize, Store Ref (1→3), Normalize (On). See Figure 2-14.
This will activate the trace 1 minus trace 3 function and display the results in trace 1. The normalized trace or flat line represents 0 dB return loss. Normalization occurs each sweep. Replace the short circuit with the DUT.

Making Complex Measurements Making a Reflection Calibration Measurement
Figure 2-14 Short Circuit Normalized

Figure 2-15 Measuring the Return Loss of the Filter


Making Complex Measurements Making a Reflection Calibration Measurement
Converting Return Loss to VSWR
Return loss can be expressed as a voltage standing wave ratio (VSWR) value using the following table or formula:
Table 2-1 Power to VSWR Conversion
4.0 4.42 14.0 1.50 18.0 1.29 28.0 1.08 38.0 1.03
6.0 3.01 14.2 1.48 18.5 1.27 28.5 1.08 38.5 1.02
8.0 2.32 14.4 1.47 19.0 1.25 29.0 1.07 39.0 1.02
10.0 1.92 14.6 1.46 19.5 1.24 29.5 1.07 39.5 1.02
10.5 1.85 14.8 1.44 20.0 1.22 30.0 1.07 40.0 1.02
11.0 1.78 15.0 1.43 20.5 1.21 30.5 1.06 40.5 1.02
11.2 1.76 15.2 1.42 21.0 1.20 31.0 1.06 41.0 1.02
11.4 1.74 15.4 1.41 21.5 1.18 31.5 1.05 41.5 1.02
11.6 1.71 15.6 1.40 22.0 1.17 32.0 1.05 42.0 1.02
11.8 1.69 15.8 1.39 22.5 1.16 32.5 1.05 42.5 1.02
12.0 1.67 16.0 1.38 23.0 1.15 33.0 1.05 43.0 1.01
12.2 1.65 16.2 1.37 23.5 1.14 33.5 1.04 43.5 1.01
12.4 1.63 16.4 1.36 24.0 1.13 34.0 1.04 44.0 1.01
12.6 1.61 16.6 1.35 24.5 1.13 34.5 1.04 44.5 1.01
12.8 1.59 16.8 1.34 25.0 1.12 35.0 1.04 45.0 1.01
13.0 1.58 17.0 1.33 25.5 1.11 35.5 1.03 45.5 1.01
13.2 1.56 17.2 1.32 26.0 1.11 36.0 1.03 46.0 1.01
13.4 1.54 17.4 1.31 26.5 1.10 36.5 1.03 46.5 1.01
13.6 1.53 17.6 1.30 27.0 1.09 37.0 1.03 47.0 1.01
13.8 1.51 17.8 1.30 27.5 1.09 37.5 1.03 47.5 1.01
VSWR 1 10
RL –
Where: RL is the measured return loss value.
VSWR is sometimes stated as a ratio. For example: 1.2:1 “one point two to one” VSWR. The first number is the VSWR value taken from the table or calculated using the formula. The second number is always 1.

Making Complex Measurements Demodulating and Listening to an AM Signal
The functions listed in the menu under Det/Demod allow you to demodulate and hear signal information displayed on the analyzer. Simply place a marker on a signal of interest, activate AM demodulation, turn the speaker on, and then listen.
Figure 2-16 Demodulation of an AM Signal


Making Complex Measurements Demodulating and Listening to an AM Signal
10.Set the signal in the top two divisions of the screen by changing the reference level. Press AMPLITUDE, and then the step down key (↓) until the signal is in the top two divisions. Set the amplitude scale to linear by pressing Scale Type (Lin). Keep the signal center displayed by pressing AMPLITUDE, Ref Level and using the (↑) (↓) step keys.

Making Complex Measurements Demodulating and Listening to an AM Signal
11.Press Det/Demod, Detector, Sample to set the detector mode of the analyzer to Sample.
12.Press Det/Demod, Demod, AM. Use the front panel volume knob to control the speaker volume.
13.You can turn your analyzer into a % AM indicator as follows:
a. Set trigger to free run by pressing Trig, Free Run.
b. Set the sweep time to 5 seconds by pressing Sweep,
Sweep Time, 5, s.
c. Set the video filter to 30 Hz by pressing BW/Avg, Video BW, 30, Hz.
d. Change the reference level to position the trace at midscreen by
pressing AMPLITUDE, Ref Level, and adjust the reference level using the front-panel knob.
e. Reset the video filter to a high value. For example, press BW/Avg,
Video BW, 30, kHz.
f. Set the sweep time to 5 milliseconds by pressing Sweep, Sweep
Time, 5, ms.
The center horizontal line of the graticule now represents 0% AM; the top and bottom lines, 100% AM. See Figure 2-17.
14.The signal to the speaker will be interrupted during retrace because the analyzer is performing automatic alignment routines. To eliminate the interruption and clicks between sweeps, turn the auto alignment function off by pressing System, Alignments, Auto Align, Off.

Making Complex Measurements Demodulating and Listening to an AM Signal
Figure 2-17 Continuous Demodulation of an AM Signal


Making Complex Measurements Demodulating and Listening to an AM Signal