DUAL OUTPUT POWER SUPPLY

Agilent MODEL E3630A

OPERATING AND SERVICE MANUAL FOR INSTRUMENTS WITH SERIAL NUMBERS

KR85014528 AND ABOVE

For instruments with Serial Numbers above KR85014528, a change page may be included.

Manual Part No. 5959-5329

April 2000 · Edition 7

Page 2 of the Agilent Technologies E3630A user manual — NOTE
Page 2 · click to enlarge
Page 2

SAFETY SUMMARY

The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies assumes no liability for the customer's failure to comply with these requirements.

SAFETY SYMBOLS

BEFORE APPLYING POWER. Verify that the product is set to match the available line volt- age and that the correct fuse is installed.

Instruction manual symbol; the product will be marked with this symbol when it is necessary for the user to refer to the instruction manual.

!

GROUND THE INSTRUMENT. This product is a Safety Class I instrument (provided with a protective earth terminal). To minimize shock hazard, the instrument chassis and cabinet must be connected to an electrical ground. The instrument must be connected to the ac power supply mains through a three-conductor power cable, with the third wire firmly connected to an electrical ground(safety ground) at the power outlet. Any interruption of the protective(grounding) conductor or disconnection of the protective earth terminal will cause a potential shock hazard that could result in personal injury. If the instrument is to be energized via an external autotransformer for voltage reduc- tion, be certain that the autotransformer common terminal is connected to the neutral(earthed pole) of the ac power lines (supply mains).

Indicate earth(ground) terminal.

The WARNING sign denotes a hazard. It calls attention to a procedure, practice, or the like, which, if not correctly per- formed or adhered to, could result inper- sonal injury. Do not proceed beyond a WARNING sign until the indicated con- ditions are fully understood and met.

WARNING

The CAUTION sign denotes a hazard. It calls attention to an operating procedure, or the like, which, if not correctly per- formed or adhered to, could result in damage to or destruction of part or all of the product. Do not proceed beyond CAUTION sign until the indicated condi- tions are fully understood and met.

DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE. Do not operate the instrument in the presence of flammable gases or fumes.

CAUTION

KEEP AWAY FROM LIVE CIRCUITS. Operating personnel must not remove instrument covers. Component replacement and internal adjustments must be made by qualified service personnel. Do not replace compo- nents with power cable connected. Under certain conditions, dangerous voltages may exist even with the power cable removed. To avoid injuries, always disconnect power, dis- charge circuits and remove external voltage sources before touching components.

The NOTE sign denotes important infor- mation. It calls attention to a procedure, practice, condition or the like, which is essential to highlight.

NOTE

DO NOT SERVICE OR ADJUST ALONE. Do not attempt internal service or adjustment unless another person, capable of rendering first aid and resuscitation, is present.

DO NOT SUBSTITUTE PARTS OR MODIFY INSTRUMENT. Because of the danger of introducing additional hazards, do not install substitute parts or perform any unauthorized modi- fication to the instrument. Return the instrument to a Agilent Technologies Sales and Service Office for service and repair to ensure that safety features are maintained.

Instruments that appear damaged or defective should be made inoperative and secured against unintended operation until they can be repaired by qualified service personnel.

or

Page 3 of the Agilent Technologies E3630A user manual — Table of Contents
Page 3 · click to enlarge
Page 3

Table of Contents

SAFETY SUMMARY 1-2

GENERAL INFORMATION 1-4

INTRODUCTION 1-4 SAFETY CONSIDERATIONS 1-4 SAFETY AND EMC REQUIREMENTS 1-4 INSTRUMENT AND MANUAL IDENTIFICATION 1-4 OPTIONS 1-4 ACCESSORY 1-4 DESCRIPTION 1-4 LINE FUSE 1-5 SPECIFICATIONS 1-5

INSTALLATION 1-6

INITIAL INSPECTION 1-6

Mechanical Check 1-6 Electrical Check 1-6 INSTALLATION DATA 1-6

Location and Cooling 1-6 Outline Diagram 1-6 Rack Mounting 1-6 INPUT POWER REQUIREMENTS 1-6

Power Cable 1-6

OPERATING INSTRUCTIONS 1-6

INTRODUCTION 1-6 TURN -ON CHECKOUT PROCEDURE 1-7 OPERATION 1-7

Tracking Ratio Control 1-7 Overload Protection Circuits 1-7 Operation Beyond Rated Output 1-7 Connecting Load 1-8 Parallel Operation 1-8 Series Operation 1-8

LOAD CONSIDERATIONS 1-8

PULSE LOADING 1-8 REVERSE CURRENT LOADING 1-8 OUTPUT CAPACITANCE 1-8 REVERSE VOLTAGE PROTECTION 1-8

Page 4 of the Agilent Technologies E3630A user manual — GENERAL INFORMATION
Page 4 · click to enlarge
Page 4

GENERAL INFORMATION

between your instrument and the instrument described by this manual. The Change Sheet may also contain information for correcting errors in the manual.

INTRODUCTION This section contains general information concerning the E3630A triple output power supply. Included are safety con- siderations, safety and EMC requirements, instrument and manual identification, option and accessory information, instrument description, and specifications.

OPTIONS Options OE3 and OE9 determine which line voltage is selected at the factory. The standard unit is configured for 115 Vac ± 10%, 47-63 Hz input.

SAFETY CONSIDERATIONS This product is a Safety Class I instrument, which means that it is provided with a protective earth ground terminal. This ter- minal must be connected to an ac source that has a 3-wire ground receptacle. Review the instrument rear panel and this manual for safety markings and instructions before operating the instrument. Refer to the Safety Summary page at the beginning of this manual for a summary of general safety information. Specific safety information is located at the appropriate places in this manual.

Option No. Description OE3: 230 Vac ± 10%, 47-63 Hz Input OE9: 100 Vac ± 10%, 47-63 Hz Input 910: One additional operating and service manual shipped with the power supply

ACCESSORY The accessory listed below may be ordered from your local Agilent Technologies Sales Office either with the power sup- ply or separately. (Refer to the list at the rear of the manual for address.)

SAFETY AND EMC REQUIREMENTS This power supply is designed to comply with the following safety and EMC(Electromagnetic Compatibility) require- ments:

Agilent Part No.Description 5063-9767 Rack Kit for mounting one or two 3 1/2" high

supplies in a standard 19" rack

IEC 1010-1(1990)/EN 61010 (1993): Safety Require- ments for Electrical Equipment for Measurement, Control, and Laboratory Use CSA C22.2 No.231: Safety Requirements for Electrical and Electronic Measuring and Test Equipment UL 1244: Electrical and Electronic Measuring and Testing Equipment EMC Directive 89/336/EEC: Council Directive entitled Approximation of the Laws of the Member States relating to Electromagnetic Compatibility EN 55011(1991) Group 1, Class B/CISPR 11 (1990): Lim- its and Methods of Radio Interference Characteristics of Industrial, Scientific, and Medical(ISM) Radio - Fre- quency Equipment EN 50082-1(1992) /

The rack mount kit is needed for rack mounting of the E3630A power supply.

DESCRIPTION This constant-voltage/current-limiting triple output power sup- ply combines two 0 to ±20V tracking outputs rated at 0.5 amps with an additional single output that is rated at 0 to 6 volts and 2.5 amps. The +20V and -20V tracking outputs can also be used in series as a single 0 to 40V 0.5-amp output. Connections to the supply's output and to chassis ground are made to binding posts on the front panel. The supply's three outputs share a common output terminal, which is isolated from chassis ground so that any one output terminal can be grounded.

IEC 801-2(1991)Electrostatic Discharge Require-

All outputs are protected against overload and short-circuit damage. The ±20V outputs are protected by circuits that limit the output current to 110% of its nominal maximum. The over- load protection circuit for the +6V output has a current fold- back characteristic that reduces the output current as an overload increases until only 1 amp flows through a short cir- cuit. The 6V output current limit depends on the output termi- nal voltage and varies linearly between 2.75 amps at 6 volts and 1 amp at zero volts.

ments IEC 801-3(1984)Radiated Electromagnetic Field
Requirements IEC 801-4(1988)Electrical Fast Transient/Burst

INSTRUMENT AND MANUAL IDENTIFICATION A serial number identifies your power supply. The serial num- ber encodes the country of manufacture, the week of the lat- est significant design change, and a unique sequential number. The letter "KR" designates Korea as the country of manufacture, the first one digit indicates the year (3=1993, 4=1994, and so forth), and the second two digits indicate the week. The remaining digits of the serial number are a unique, five-digit number assigned sequentially.

All controls, digital meter, and output terminals are located on the front panel. One voltage control sets the 0 to 6V and another sets the voltages of the 0 to +20V and 0 to -20V outputs simul- taneously. These dual tracking outputs are made more versatile by providing a tracking ratio control in addition to the usual volt- age control. With the tracking ratio control turned fully clockwise to its "fixed" position, the dual outputs have a fixed 1:1 tracking ratio. As the ±20V voltage control is adjusted, the voltage of the negative supply tracks the positive output within ±1%. Turning the tracking ratio control away from its fully clockwise position switches the dual tracking outputs into a variable tracking ratio

If the serial number on your supply does not agree with those on the title page of the manual, a yellow Change Sheet is supplied with the manual to explain the difference

Requirements

Page 5 of the Agilent Technologies E3630A user manual — DIMENSIONS
Page 5 · click to enlarge
Page 5

mode. In this mode the voltage of the negative output can be set lower than that of the positive output. The tracking ratio control allows the negative supply's output to be set to any value between a maximum that is within ±5% of the positive supply's output and a minimum that is less than 0.5 volts. Once a ratio is established by the tracking ratio control, the ratio of the positive output voltage to the negative output voltage remains constant as the ±20V voltage control varies the 0 to +20V output over its range.

are available for nominal inputs of 100 and 230 Vac. The sup- ply is furnished with a detachable 3-wire grounding type line cord. The ac line fuse is in an extractor type fuseholder on the rear heat sink.

LINE FUSE The line fuse is located by the ac line receptacle. Check the rating of the line fuse and replace it with the correct fuse if necessary as indicated below. These are slow-blow fuses.

The front panel also contains a line switch, three overload indicators for +6V output, +20V output, and -20V output, a voltmeter and an ammeter, and three push-button meter switches. The push buttons select one of the supply's three outputs for display. The voltmeter and ammeter always moni- tor any one supply simultaneously. In addition to the standard 115 Vac ± 10% 47 to 63 Hz input, two other line voltage options

Line Voltage Fuse Agilent Part No. 100/115 Vac 1.6 A 2110-0918 230 Vac 1.0 A 2110-0599

SPECIFICATIONS Table 1 lists detailed specifications for the power supply.

Table 1. Specifications

AC INPUT

OPERATING TEMPERATURE RANGE

Standard: 115 Vac ± 10%, 47-63 Hz, 115 VA, 84 W OE9: 100 Vac ± 10%, 47-63 Hz, 115 VA, 84 W OE3: 230 Vac ± 10%, 47-63 Hz, 115 VA, 84 W

0 to 40 oC for full rated output. At higher temperatures, out-

put current is derated linearly to 50% at 55 oC maximum temperature.

DC OUTPUT and OVERLOAD PROTECTION

TEMPERATURE COEFFICIENT

0 to ±20 V Outputs: Maximum rated output current is 0.5 A. Short circuit output current is 0.55 A ± 5% and a fixed cur- rent limit circuit limits the output of each supply to this maxi- mum at any output voltage setting. Unbalanced loads within current rating are permitted.

All OutputsLess than 0.02% plus 1 mV voltage change

per oC over the operating range from 0 to 40 oC after 30 minutes warm-up.

STABILITY (OUTPUT DRIFT)

All Outputs: Less than 0.1% plus 5 mV (dc to 20 Hz) dur- ing 8 hours at constant line, load and ambient after an initial warm-up time of 30 minutes.

0 to +6 V Output: Maximum rated output current is 2.5 A at 6 V. The maximum available output current decreases with the output voltage setting. A current foldback circuit limits the output to 2.75 A ± 5% at 6 volts and, with decreasing voltage, reduces the current limit linearly to 1 A ± 15% at zero volts (short circuited).

LOAD TRANSIENT RESPONSE TIME

All Outputs: Less than 50 sec for output recovery to within 15 mV of nominal output voltage following a load change from full load to half load, or vice versa.

TRACKING ACCURACY

The +20 V and -20 V outputs track within 1% with the TRACKING ratio control in the Fixed position. (In variable tracking ratio mode, the negative tracking output can be adjusted from less than 0.5 V to within ±5% of the setting of the positive output.)

OUTPUT VOLTAGE OVERSHOOT

All Outputs: During turn-on or turn-off of ac power, output plus overshoot will not exceed 1 V if the out- put control is set for less than 1 V. If the con- trol is set for 1 V or higher, there is no overshoot.

LOAD REGULATION

All Outputs: Less than 0.01% plus 2 mV for a full load to no load change in output current.

METER ACCURACY±(0.5% of output + 2 counts) at

25oC±5oC

LINE REGULATION

All Outputs: Less than 0.01% plus 2 mV for any line volt- age change within rating.

METER RESOLUTION

All OutputVoltage 10 mV Current 10 mA

PARD (Ripple and Noise)

All Outputs: Less than 0.35 mV rms/1.5 mV p-p (20 Hz-20 MHz). Common Mode Current (CMI): Less than 1 A for all

DIMENSIONS

212.3 mmW x 88.1 mmH x 269.2 mmD (8.354 inW x 3.469 inH x 10.591 inD)

outputs (20 Hz-20 MHz).

WEIGHT3.8 kg(8.4 lbs) net, 5.1 kg(11.3 lbs) shipping
Page 6 of the Agilent Technologies E3630A user manual — OPERATING INSTRUCTIONS
Page 6 · click to enlarge
Page 6

INSTALLATION

cord plug appropriate for the user's location. Notify the near- est Agilent Sales and Service Office if the appropriate power cord is not included with the supply.

INITIAL INSPECTION Before shipment, this instrument was inspected and found to be free of mechanical and electrical defects. As soon as the instrument is unpacked, inspect for any damage that may have occurred in transit. Save all packing materials until the inspection is completed. If damage is found, a claim should be filed with the carrier. The Agilent Technologies Sales and Service office should be notified as soon as possible.

Agilent Technologies E3630A — INSTALLATION diagram (page 6)
Fig. 1

Mechanical Check This check should confirm that there are no broken knobs or connectors, that the cabinet and panel surfaces are free of dents and scratches, and that the meter is not scratched or cracked.

Electrical Check Perform the TURN-ON CHECKOUT PROCEDURE in the fol- lowing paragraph to confirm that the supply is operational. Alternately, check the supply more fully using the PERFOR- MANCE TEST in the service information section.

INSTALLATION DATA The instrument is shipped ready for bench operation. Before applying power to the supply, please read the INPUT POWER REQUIREMENTS paragraph.

Figure 1. Outline Diagram

OPERATING INSTRUCTIONS

Location and Cooling This instrument is air cooled. Sufficient space should be allowed so that a free flow of cooling air can reach the sides and rear of the instrument when it is in operation. It should be used in an area where the ambient temperature does not

INTRODUCTION The following steps describe the use of the front panel con- trols and indicators illustrated in Figure 2 and serve as a brief check that the supply is operational. Follow this checkout pro- cedure or the more detailed performance test of service infor- mation section when the instrument is received and before it is connected to any load equipment.

exceed 40oC.

Outline Diagram Figure 1 illustrates the outline shape and dimensions of the supply.

Agilent Technologies E3630A — OPERATING INSTRUCTIONS diagram (page 6)
Fig. 2

Rack Mounting This supply may be rack mounted in a standard 19-inch rack panel either by itself or alongside a similar unit. Please see the ACCESSORY, page 1-4, for available rack mounting accessory. The rack-mounting kit includes complete installa- tion instructions.

INPUT POWER REQUIREMENTS Depending on the line voltage option ordered, the supply is ready to be operated from one of the power sources listed in Table 1. A label on the rear heat sink shows the nominal input voltage set for the supply at the factory.

Power Cable To protect operating personnel, the supply should be grounded. This supply is equipped with a three conductor power cable. The third conductor is the ground conductor and when the cable is plugged into an appropriate receptacle, the supply is grounded.

The power supply is equipped at the factory with a power

Figure 2. Front-Panel Controls and Indicators

Page 7 of the Agilent Technologies E3630A user manual — TURN -ON CHECKOUT PROCEDURE
Page 7 · click to enlarge
Page 7

tor the 0 to 40V output voltage, add the voltmeter readings of the +20V and -20V output and use either the +20V or the - 20V meter to measure the current.

Before applying power to the supply, check the label on the heat sink to make certain that the supply's line voltage option agrees with the line voltage to be used. If the option does not correspond to your line voltage, refer to paragraph "LINE VOLTAGE OPTION CONVERSION" in the service section before applying power.

Tracking Ratio Control With the Tracking Ratio control in the Fixed position, the volt- age of the -20V supply tracks that of the +20V supply within 1% for convenience in varying the symmetrical voltages needed by operational amplifiers and other circuits using bal- anced positive and negative inputs.

TURN -ON CHECKOUT PROCEDURE

Turn the Tracking Ratio control counter clockwise out of the Fixed position to set the voltage of the -20V supply lower than that of the +20V supply. The negative supply can be set from a minimum of less than 0.5 volts to a maximum within 5% of the +20V supply's output. Once this is done, the ±20V voltage control still controls both outputs and maintains a constant ratio between their voltages.

a. Connect line cord to power source and turn LINE switch on.

b. Push +6V METER switch and, with no load con- nected, vary +6V VOLTAGE control over its range and check that the voltmeter responds to the control setting and the ammeter indicates zero.

c. Set the +6V VOLTAGE control to 6 volts and short the +6V output terminal to COM (common) terminal with an insulated test lead. The ammeter should indicate a short-circuit output current of approximately 1.0 A. Remove the short from the output terminals.

±20-Volt Current Limit. The +20V and -20V outputs are indi- vidually protected against overload or short circuit damage by separate current limit circuits to limit the output current to 0.55 A ± 5%. (This is 110% of the maximum rated output.) If a sin- gle load is connected between the +20V and -20V outputs, the circuit set for the lesser current limit will limit the output. No deterioration of supply performance occurs if the output current remains below the current limit setting.

d. Push the +20V METER switch and turn Tracking Ratio control fully clockwise to the Fixed position. With no load connected, vary ±20V VOLTAGE control

over its range and check that the voltmeter responds to the control setting and the ammeter indi- cates zero.

e. Set the ±20V VOLTAGE control to 20 volts and short the +20V output terminal to the COM terminal with an insulated test lead. The ammeter should indicate a short-circuit output current of 0.55 A ± 5%. Remove the short from the output terminals.

+6V Current Foldback. The overload and short-circuit pro- tection circuit for the +6V output reduces the output current limit as the output terminal voltage decreases. (The operating region of the +6V output is enclosed by heavy lines in Figure 4.) The maximum rated output current is 2.5 A and the current limit is factory-adjusted to operate at 2.75 A ± 5% when the output is 6 volts. At lower output voltages, the circuit reduces the maximum obtainable output current linearly until 1 A ± 15% flows when the output is shorted. The short-circuit cur- rent can not be adjusted.

f. Repeat steps (d) and (e) for -20 V output.

g. Adjust the +20V output to 20 volts. Then push -20V METER switch and check the effect of the Tracking Ratio control on the voltage of the -20V output. The - 20V output should be adjustable from less than 0.5 volts to a maximum of 19 to 21 volts.

If this brief checkout procedure or later use of the supply reveals a possible malfunction, see the service information section for detailed test, troubleshooting, and adjustment pro- cedures.

During the actual operation of the ±20V and +6V outputs, if a load change causes the current limit to be exceeded, the OL LED is lighted. If overload conditions occur, the ±20V supplies will protect the load by limiting the current to 0.55 A and the +6V supply will protect the load by reducing both voltage and current simultaneously along the foldback locus as shown in Figure 4. The ±20V and +6V supplies are self restoring; that is, when the overload is removed or corrected, the output voltage is automatically restored to the previously set value.

OPERATION This power supply can be operated individually or in parallel or series with another supply (see Parallel and Series Opera- tion paragraphs). All output terminals are isolated from ground. The ±20V and +6V outputs use a single common out- put terminal. This common (COM) terminal or any one of the other output terminals may be grounded to the chassis at the front panel ground terminal ( in Figure 2), or all outputs may be left floating. Loads can be connected separately between each of the 0 to ±20V output terminals and the COM terminal, or between the -20V and the +20V terminals for a 0 to 40V output. Each output voltage or current can be quickly selected for monitoring with the push-button meter switches. To moni-

Operation Beyond Rated Output The supply may be able to provide voltages and currents greater than its rated maximum outputs if the line voltage is at or above its nominal value. Operation can be extended up to 5% over the rated output without damage to the supply, but performance can not be guaranteed to meet specifications in

Overload Protection Circuits

Page 8 of the Agilent Technologies E3630A user manual — LOAD CONSIDERATIONS
Page 8 · click to enlarge
Page 8

necessary to fulfill the total current demand.

Series Operation Series operation of two or more power supplies can be accomplished up to the output isolation rating of any one sup- ply to obtain a higher voltage than that available from a single supply. Series connected supplies can be operated with one load across both supplies or with a separate load for each supply. The power supply has a reverse polarity diode con- nected across the output terminals so that if operated in series with other supplies, damage will not occur if the load is short-circuited or if one supply is turned on separately from its series partners. When this connection is used, the output volt- age is the sum of the voltages of the individual supplies. Each of the individual supplies must be adjusted in order to obtain the total output voltage.

LOAD CONSIDERATIONS

This section provides information on operating your supply with various types of loads connected to its output.

PULSE LOADING The power supply will automatically cross over from constant- voltage to current-limit operation in response to an increase in the output current over the preset limit. Although the preset limit may be set higher than the average output current, high peak currents (as occur in pulse loading) may exceed the pre- set current limit and cause crossover to occur and degrade performance.

Figure 3. Current Limit Characteristic of the 6V Supply

this region. If the line voltage is maintained in the upper end of the input voltage range, however, the supply probably will operate within its specifications

Connecting Load Each load should be connected to the power supply output terminals using separate pairs of connecting wires. This will minimize mutual coupling effects between loads and takes full advantage of the low output impedance of the supply. Load wires must be of adequately heavy gauge to maintain satis- factory regulation at the load.

REVERSE CURRENT LOADING An active load connected to the supply may actually deliver a reverse current to the supply during a portion of its operating cycle. An external source can not be allowed to pump current into the supply without risking loss of regulation and possible damage to the output capacitor of the supply. To avoid these effects, it is necessary to preload the supply with a dummy load resistor so that the supply delivers current through the entire operating cycle of the load devices.

Each pair of connecting wires should be as short as possible and twisted or shielded to reduce noise pick-up. If a shield is used, connect one end to the supply ground terminal and leave the other end unconnected.

OUTPUT CAPACITANCE An internal capacitor across the output terminals of the supply helps to supply high-current pulses of short duration during constant-voltage operation. Any capacitance added externally will improve the pulse current capability, but will decrease the load protection provided by the current limiting circuit. A high- current pulse may damage load components before the aver- age output current is large enough to cause the current limit- ing circuit to operate.

If load considerations require locating output power distribu- tion terminals at a distance from the power supply, then the power supply output terminals should be connected to the remote distribution terminals by a pair of twisted or shielded wires and each load should be connected to the remote distri- bution terminals separately.

Parallel Operation Two or more supplies can be connected in parallel to obtain a total output current greater than that available from one sup- ply. The total output current is the sum of the output currents of the individual supplies. The output voltage controls of one power supply should be set to the desired output voltage, and the other supply set for a slightly larger output voltage. The supply set to the lower output voltage will act as a constant voltage source, while the supply set to the higher output will act as a current-limited source, dropping its output voltage until it equals that of the other supply. The constant voltage source will deliver only that fraction of its rated output current

REVERSE VOLTAGE PROTECTION A diode is connected across the output terminals with reverse polarity. This diode protects the output electrolytic capacitors and the series regulator transistors from the effects of a reverse voltage applied across the output terminals. Since series regulator transistors can not withstand reverse voltage either, diodes are also connected across them. When operat- ing supplies in parallel, these diodes protect an unenergized supply that is in parallel with an energized supply.

Page 9 of the Agilent Technologies E3630A user manual — Load Considerations
Page 9 · click to enlarge
Page 9

Figure 4. Reverse Current Loading Solution

Page 10 of the Agilent Technologies E3630A user manual — PRINCIPLES OF OPERATION (Block Diagram Overview)
Page 10 · click to enlarge
Page 10

SERVICE INFORMATION

Figure A-1. Block Diagram

PRINCIPLES OF OPERATION (Block Diagram Overview)

By comparing its output to a high-stability reference, the 0 to +6-volt regulator holds its output voltage at the value deter- mined by a front panel control. Any error in the actual output as compared to the desired output is amplified by an opera- tional amplifier and applied as feedback to control the con- duction of a series regulator transistor. As a result, the voltage across the series transistor varies so as to hold the output voltage constant at the desired level. The high gain of the voltage comparison amplifier and the stability of the reference voltage ensure that input voltage or load current variations have little effect on the output voltage.

This section presents the principles of operation of the E3630A Triple Output Power Supply. Throughout this discus- sion, refer to both the block diagram of Figure A-1 and the schematic diagrams at the rear of the manual.

The two primary windings of the power transformer are con- nected in one of three different ways by setting the two slide switches mounted on the circuit board. These switches select one of the nominal ac input voltages for which the supply is designed: 100 V, 115 V, or 230 V.

The 0 to +6-volt output is protected by a current foldback lim- iter to minimize dissipation in the series regulator transistor during overloads. In a current foldback circuit, the current limit depends on the output terminal voltage. The current limit ranges from 2.75 A ± 5% at 6 volts to 1 A ± 15% with the out- put shorted. The operating region of the +6-volt regulator out-

The transformer secondaries, together with rectifiers and capacitor filters, provide raw dc for the three output regulator circuits and for another regulator that provides reference and bias voltages to the output regulators.

Page 11 of the Agilent Technologies E3630A user manual — MAINTENANCE
Page 11 · click to enlarge
Page 11

put is enclosed by a heavy line in Figure 3 of the operating section. If the operating point reaches the diagonal current limit line, a decrease in load resistance moves the operating point down the line, reducing the output voltage and current. Current foldback is controlled by a second operational ampli- fier (current comparison amplifier) in the regulator that moni- tors the dc output current. This current comparison amplifier takes control of the output away from the voltage comparison amplifier when the current reaches the design limit. Removing the overload restores constant voltage operation auto- matically.

The reference and bias supply powers the operation amplifi- ers and provides reference and bias voltages for the output regulators. The display power circuit provides voltage which is used by the A/D converter and display.

The turn-on/turn-off control circuit prevents output transients when the supply is turned on or off. It does this by delaying the application of certain bias and reference voltages at turn- on and removing them shortly after turn-off.

Three meter push-button switches select which of the sup- plies has its output voltage and current indicated on the front panel meters.

The 0 to +20-volt regulator has a fixed current limit at 105% of its 0.5 amp maximum rated output. The input ac line voltage is first applied to the preregulator which operates in conjunc- tion with the SCR control circuit (preregulator control circuit) to rectify the tap switched AC voltage. This preregulator mini- mizes the power dissipated in the series regulating elements by controlling the dc level across the input filter capacitor, depending on the output voltage. To achieve this, tap switch- ing is accomplished by two SCRs and one bridge diode (CR28, CR32 and CR26) and the SCR control circuit. This cir- cuit allows the input capacitor to charge to one of two discrete voltage levels depending on the output required.

Diode CR2, CR3, and CR4 are connected across the output terminals in reverse polarity. They protect the output elec- trolytic capacitor and the series regulator transistors from a reverse voltage applied across the output terminals.

MAINTENANCE

INTRODUCTION This section provides performance test, troubleshooting infor- mation, and adjustment and calibration procedures. The fol- lowing operation verification tests comprise a short procedure to verify that the power supply is performing properly, without testing all specified parameters.

When output voltage exceeds the reference level, the SCR control circuit fires two SCRs that cause the input capacitor to be charged to the voltage which is necessary for full output of the supply. When the two SCRs are not fired, the bridge diode CR26 conducts and half the voltage is applied to series pass transistor Q9.

If a fault is detected in the power supply while making the performance check or during normal operation, proceed to the troubleshooting procedures. After troubleshooting, per- form any necessary adjustments and calibrations. Before returning the power supply to normal operation, repeat the performance check to ensure that the fault has been properly corrected and that no other faults exist.

The 0 to -20-volt regulator is, in turn, similar to the +20-volt regulator except that it resembles a complementary mirror image of the latter. The output voltages of the +20-volt and - 20-volt supplies are both set by the same front panel control and track each other within 1% in the fixed tracking ratio mode. Precise tracking of the two outputs is achieved by controlling the positive output conventionally and using that output as the reference voltage for the negative output.

Test Equipment Required The following Table A-1 lists the equipment required to per- form the various procedures described in this section.

Table A-1. Test Equipment Required

TYPEREQUIRED CHARACTERISTICSUSERECOMMENDED MODEL OscilloscopeSensitivity : 100 V Bandwidth : 20 MHz/100 MHzDisplay transient response and ripple and noise waveforms.Agilent 54600A RMS VoltmeterTrue rms, 20 MHz bandwidth Sensitivity : 1 mV Accuracy : 5%Measure rms ripple and noise voltage. MultimeterResolution : 100 nV Accuracy : 0.0035%Measure dc voltages.Agilent 34401A Electronic LoadVoltage Range : 240 Vdc Current Range : 10 Adc Open and short switches Transient on/offMeasure load and line regulation.Agilent 6063A Resistive Loads (R ) L40 20 W, 2.4 20 WMeasure ripple and noise.
Page 12 of the Agilent Technologies E3630A user manual — Figure A-2. Line Voltage Selector (set for 115 Vac)
Page 12 · click to enlarge
Page 12

Table A-1. Test Equipment Required (Cont’d)

Current Sampling 1 Resistor (Shunt)00 mΩ 0.1% 15 WMeasure output current Variable Voltage R Auto Transformer Vange : 85-130 and 200-260 oltsVary ac input.

Operation Verification Tests The following tests assure that the power supply is per- forming properly. They do not, however, check all the speci- fied parameters tested in the complete performance test described below. Proceed as follows: a. Perform turn-on checkout procedure given in page 1-7. b. Perform the load regulation performance tests given in the following paragraphs.

General Measurement Techniques

Connecting Measuring Devices. To achieve valid results when measuring load regulation, ripple and noise, and tran- sient response time of the supply, measuring devices must be connected as close to the output terminals as possible. A measurement made across the load includes the impedance of the leads to the load. The impedance of the load leads can easily be several orders of the magnitude greater than the supply impedance and thus invalidate the measurement. To avoid mutual coupling effects, each measuring device must be connected directly to the output terminals by separate pairs of leads.

Line Voltage Option Conversion To convert the supply from one line voltage option to another, the following three steps are necessary: a. After making certain that the line cord is disconnected from a source of power, remove the top cover from the supply and set the two sections of the line voltage selec- tor switch for the desired line voltage (see Figure A-2). b. Check the rating of the installed fuse and replace it with the correct value, if necessary. For Option OE3, use a slow-blow 1.0-amp fuse. For standard and Option OE9, use a slow-blow 1.6-amp fuse. c. Mark the instrument clearly with a tag or label indicating the correct line voltage to be used.

When performance measurements are made at the front ter- minals (Figure A-3) the load should be plugged into the front of the terminals at (B) while the monitoring device is con- nected to a small lead or bus wire inserted through the hole in the neck of the binding post at (A). Connecting the measuring device at (B) would result in a measurement that includes the resistance of the leads between the output terminals and the point of connection.

Figure A-2. Line Voltage Selector (set for 115 Vac)

Figure A-3. Front Panel Terminal Connections

PERFORMANCE TESTS The following paragraphs provide test procedures for verify- ing the power supply's compliance with the specifications of Table 1. Proceed to the troubleshooting procedures if you observe any out of specification performance.

Selecting Load Resistors. Power supply specifications are checked with a full load resistance connected across the sup- ply output. The resistance and wattage of the load resistor, therefore, must permit operation of the supply at its rated out- put voltage and current. For example, a supply rated at 20 volts and 0.5 amperes would require a load resistance of 40 Ω at the rated output voltage. The wattage rating of this resistor would have to be at least 20 watts.

Electronic Load. Some of the performance test procedures use an electronic load to test the supply quickly and accu- rately. An electronic load is considerably easier to use than a load resistor. It eliminates the need for connecting resistors or rheostats in parallel to handle the power, it is much more sta- ble than a carbon-pile load. It is easier to switch between load conditions as required for the load regulation and load tran- sient response tests.

Before applying power to the supply, make certain that its line voltage selector switch (S2) is set for the line voltage to be used. (See CAUTION notice in operating section for additional information on S2.)

Page 13 of the Agilent Technologies E3630A user manual — Service Information
Page 13 · click to enlarge
Page 13

Output Current Measurement. For accurate output current measurements, a current sampling resistor should be inserted between the load and the output of the supply. To simplify grounding problems, one end of this sampling resistor should be connected to the same output terminal of the supply which will be shorted to ground. An accurate voltmeter is then placed across the sampling resistor and the output current calculated by dividing the voltage across the sampling resistor by its ohmic value. The total resistance of the series combination should be equal to the full load resistance as determined in the preceding paragraphs. Of course, if the value of the sampling resistor is very low when compared to the full load resistance, the value of the sampling resistor may be ignored. The meter shunt recom- mended in Table A-1, for example, has a resistance of only 100 m and can be neglected when calculating the load resistance of the supply. Figure A-4 shows a four terminal meter shunt. The load current through a shunt must be fed from the extremes of the wire leading to the resistor while the sampling connections are made as close as possible to the resistance portion itself.

Variable Tracking Ratio e. Leave the ±20V VOLTAGE control set as in step (d), and use a DVM to monitor the voltage of the -20V supply while adjusting the TRACKING RATIO control over its VARIABLE range. The -20V supply should be capable of being adjusted from less than 0.5 volts to between 19 to 21 volts. Return the TRACKING RATIO control to the FIXED position.

Leave the TRACKING RATIO control in the FIXED position throughout the reminder of the performance test.

Rated Output and Ammeter Accuracy

f. Connect two 40 20 W load resistors across both of the 20V outputs of the supply and set the ±20V VOLTAGE control for ±20V outputs. (All supplies must be fully loaded while checking the rated output voltage and cur- rent of each supply.) g. Connect the test setup shown Figure A-5 to the +6V output. Make the total resistance of RL and the current sampling resistor equal to 2.4 to permit operating the output at full load. RL should have a power rating of at least 20 watts. h. Close the switch and set the +6V VOLTAGE control so that the DVM indicates a voltage drop across the current sam- pling resistor that corresponds to a current of 2.5 amps. i. Push the +6V METER switch and verify that the front panel ammeter indication is within ±(0.5%+2 counts) of 2.5 amps (2.47A to 2.53A). j. Check the rated output and ammeter accuracy of the +20V and -20V supplies similarly by connecting the test setup of Figure A-5 to each output in turn. For each 20V supply: make the total resistance of RL and the current sampling resistor 40 , set the ±20V VOLTAGE control for a current indication on the DVM of 0.5 A, check that the panel meter indication is within ±(0.5%+2 counts) of 0.5 A (0.48A to 0.52A).

Figure A-4. Current Sampling Resistor Connections

Rated Output, Tracking, Meter Accuracy, and Cur- rent Limit To check that all supplies will furnish their maximum rated output voltage and current, that the ±20V outputs track each other, that the front panel meters are accurate, and that the current limit circuits function, proceed as follows:

Voltmeter Accuracy a. With no loads connected: turn on the supply, connect a digital voltmeter between the +20V terminal and common (COM), and set the ±20V VOLTAGE control so that the DVM indication is as near as possible to 17 volts. b. Push the +20V METER switch on and check the front panel voltmeter indication. It should be within ±(0.5% +2 counts) of the DVM indication (16.90V to 17.10V). c. Set the TRACKING RATIO control to the FIXED position, and check the +20V and -20V ranges of the front panel voltmeter similarly by connecting the DVM to each of these outputs in turn, setting the ±20V VOLTAGE control for a 20 volts DVM indication, and verifying that the panel meter is accurate within ±(0.5%+ 2 counts) (19.7V to 20.3V).

Tracking d. Connect the DVM to the +20V output, set the ±20V VOLTAGE control for a DVM indication of 20 volts, and reconnect the DVM to the -20V output without disturbing the voltage control. The voltage at the -20V output should be within 1% of the +20V output (19.8V to 20.2V).

Figure A-5. Output Current, Test Set UP

Page 14 of the Agilent Technologies E3630A user manual — Figure A-6. Basic Test Setup
Page 14 · click to enlarge
Page 14

Current Limit k. Disconnect all loads from the supply. l. Connect the test setup shown in Figure A-5 to the +20- volt output. Substitute a short for RL and leave the load circuit switch open. m. Set the voltage of the ±20V supplies to 20 volts. n. Close the load switch and determine the current flow through the current sampling resistor (meter shunt) by measuring its voltage drop with the DVM. The current should be 0.55A±5% (0.5225A to 0.5775A). o. Check the current limit of the -20V supply in the same way. Its short-circuit current should also be 0.55A±5% (0.5225A to 0.5775A). p. Connect the test setup shown in Figure A-5 to the +6V output. Close the switch, set the total resistance of RL and the current sampling resistor to an initial value of 2.4 Ω or greater, and set the output voltage to 6 volts. q. Reduce the value of RL gradually while observing the out- put current indicated by the DVM. The current should increase to a maximum of 2.75A±5% (2.6125A to 2.8875A) before it begins to decrease. r. Connect a short across RL and then recheck the current indicated by the DVM. The short-circuit current of this out- put should be 1A±15% (0.85A to 1.15A). Disconnect the test setup from the supply.

Line Regulation (Source Effect) Definition: The change, EOUT, in the static value of dc output voltage resulting from a change in ac input voltage from a minimum to a maximum value (±10% of nominal voltage).

To check the line regulation: a. Connect a variable autotransformer between the input power source and the power supply line plug. b. Connect the test equipment across the output of the +20V supply as shown in Figure A-6. Operate the elec- tronic load in constant current mode and set its current to the full rated value of the +20V supply. c. Adjust the autotransformer for a low line input (-10% of nominal voltage). d. Turn on the power, adjust the output of the supply to its maximum rated voltage, and record the DVM indication. e. Adjust the autotransformer for high line voltage input (+10% of nominal voltage) and recheck the DVM indica- tion. It should be within 0.01% plus 2mV of the reading in step (d). f. Repeat steps (b) through (e) for each of the remaining supply outputs.

Ripple and Noise Definition: Ripple and noise are measured in the rms or peak- to-peak value over a 20 Hz to 20 MHz bandwidth. Fluctuations below the lower frequency limit are treated as drift.

Load Regulation (Load Effect) Definition: The change, EOUT, in the static value of dc out- put voltage resulting from a change in load resistance from open circuit to the value that yields maximum rated output current (or vice versa).

RMS Measurement The rms measurement is not an ideal representation of the noise, since fairly high output noise spikes of short duration could be present in the ripple and not appreciably increase the rms value.

To check the load regulation: a. Connect the test equipment across the output of the +20V supply as shown in Figure A-6. Operate the elec- tronic load in constant current mode and set its current to the full rated value of the +20V supply. b. Turn on the supply and adjust its voltage to its maximum rated value. c. Record the voltage indicated on the DVM. d. Operate the electronic load in open (input off) mode and recheck the DVM indication after reading settles. It should be within 0.01% plus 2mV of the reading in step (c). e. Repeat steps (a) through (d) for each of the remaining supply outputs.

To measure the ripple and noise of the rms value on each output supply output: a. Connect the test equipment across the output of the +20V supply as shown in Figure A-7. b. Turn on the supply and push +20V METER switch. c. Turn up output voltage to the full rated value. d. Check that the rms noise voltage at the true rms voltme- ter is less than 0.35mV. e. Repeat for the remaining supply outputs.

Figure A-6. Basic Test Setup

Figure A-7. Ripple and Noise rms Measurement Test

Setup

·

Page 15 of the Agilent Technologies E3630A user manual — Service Information
Page 15 · click to enlarge
Page 15

Peak-to-Peak Measurement The peak-to-peak measurement is particularly important for applications where noise spikes could be detrimental to a sensitive load, such as logic circuitry.

d. Turn on the supply. e. Record the voltage across RS and convert it to current by dividing this voltage by RS. f. Check that the current is less than 1 A.

To measure the ripple and noise of the peak-to-peak value on each output supply output: a. Connect the test equipment across the output of the +20V supply as shown in Figure A-7, but replace the true rms voltmeter with the oscilloscope. b. Turn on the supply and push +20V METER switch. c. Turn up output voltage to the full rated value. d. Set the oscilloscope to AC mode and bandwidth to 20 MHz. e. Check that the peak-to-peak noise is less than 1.5 mV. f. Repeat for the remaining supply outputs.

Load Transient Response Time Definition : This is the time for the output voltage to return to within a specified band around its voltage following a change from full load to half load or half load to full load.

To measure the load transient response time: a. Connect the test equipment across the output of the +20V supply as shown in Figure A-6, but replace the DVM with the oscilloscope. Operate the electronic load in constant current mode. b. Turn on the supply. c. Turn up output voltage to the full rated value. d. Set the electronic load to transient operation mode between one half of supply's full rated value and supply's full rated value at a 1 KHz rate with 50% duty cycle. e. Set the oscilloscope for ac coupling, internal sync and lock on either the positive or negative load transient. f. Adjust the oscilloscope to display transients as in Figure A-8. g. Check that the pulse width (t2-t1) of the transients at 15 mV from the base line is no more than 50 sec as shown. h. Repeat for the remaining supply outputs.

Common Mode Current (CMI) Definition : Common mode current is that ac current compo- nent which exists between any or all supply or output lines and chassis ground.

To measure the common mode current: a. Connect the full load for +6V output. b. Connect a 100 k resistor(RS) and a 2200 pF capacitor in parallel between common terminal(COM) and chassis ground. c. Connect the DVM across RS.

Agilent Technologies E3630A — Service Information diagram (page 15)
Fig. 3 · Service Information

Figure A-8. Load Transient Response Time Waveform

Page 16 of the Agilent Technologies E3630A user manual — SYMPTOM CHECK - PROBABLE CAUSE
Page 16 · click to enlarge
Page 16

Stability (Drift) Definition: The change in output voltage (dc to 20 Hz) for the first 8 hours following a 30-minute warm-up period with con- stant input line voltage, constant load resistance and constant ambient temperature.

Before applying power to the supply, make certain that its line voltage selector switch (S2) is set for the line voltage to be used.

To measure the stability: a. Connect the test equipment across the output of the +20V supply as shown in Figure A-6. b. Operate the electronic load in constant current mode and set its current to the full rated value of power supply. c. Turn on the supply. d. Turn up output voltage to the full rated value as read on the digital voltmeter. e. After a 30-minute warm-up, note the voltage on DVM. f. The output voltage reading should deviate less than 0.1% plus 5 mV from the reading obtained in step e over a period of 8 hours. g. Repeat for the remaining supply outputs.

Initial Troubleshooting Procedure If a malfunction is found, follow the steps below: a. Disconnect input power from the supply and remove all loads from the output. b. Table A-2 lists the symptoms and probable causes of sev- eral possible troubles. If the symptoms is one of those listed, make the recommended checks. c. If none of the symptoms of Table A-2 apply, proceed to Table A-3. This table provides an initial troubleshooting procedure that also directs you to the more detailed pro- cedures which follow it.

The numbered test points referred to in the troubleshooting procedures are identified on the circuit schematic at the rear of the manual.

TROUBLESHOOTING Before attempting to troubleshoot the power supply, ensure that the fault is with the supply and not with an associated piece of equipment. You can determine this without removing the covers from the power supply by using the appropriate portions of the "Performance Test" paragraph.

Open Fuse Troubleshooting Although transients or fatigue can cause a fuse to blow, it is a good idea to inspect the unit for obvious shorts such as dam- aged wiring, charred components, or extraneous metal parts or wire clippings in contact with circuit board conductors before replacing the fuse. The rating of the correct replace- ment fuse depends on the line voltage option of the instru- ment: for Option OE3, use a slow-blow 1.0-amp fuse and standard and Option OE9, use a slow-blow 1.6-amp fuse.

Table A-2. Miscellaneous Troubles

SYMPTOM CHECK - PROBABLE CAUSE

High ripple a. Check operating setup for ground loops. b. Check main rectifiers(CR1, CR25, CR26) for open. c. Supply may be operating in current limit mode. Check current limit adjustment, steps (k)

thru (r) on page A-5.

Will not current limit Check for open OR-gate diodes (CR7, CR11, CR18) or defective current limit amplifier (U2, U6, U8).

Poor load and line regulation a. Check bias and reference voltages, Table A-4. b. Check main rectifiers and filters for opens.

Oscillation or poor transient response time

a. High frequency oscillations (above 50 kHz) can be caused by an open C11, C20, or C25. b. A defective output capacitor (C2, C3, or C4) can cause oscillations in one of many frequency

ranges. c. Oscillation only in the current limiting mode can be caused by an open C12, C19, or C24.

Transient voltage overshoot at turn-on or turn-off

a. Overshoot only in the -20V supply can be caused by a shorted Q3. b. Overshoot in all three supply outputs can be caused by an open Q2 or a shorted Q6.

Excessive heat a. Check preregulator control circuit. Refer to Table A-8 and Table A-9. b. Check CR27, CR28, CR31, CR32 for short.

Output Voltage clamped above 10V for ± 20V output

Check preregulator control circuit. Refer to Table A-8 and Table A-9.

Page 17 of the Agilent Technologies E3630A user manual — SYMPTOM STEP - ACTION RESPONSE PROBABLE CAUSE
Page 17 · click to enlarge
Page 17

Table A-3. Initial Trobleshooting Procedure

STEP ACTION RESPONSE NEXT ACTION

1 Check output voltage of +20V supply.

a. Normal b. Zero volts

a. Proceed to step (2). b. Check ac line fuse (F1). If blown, proceed to "Open Fuse

Troubleshooting" paragraph. If not blown, check bias and reference voltages (Table A-4). c. Check bias and reference voltages (Table A-4).

c. Output voltage lower or

2 Check output voltage of -20V supply in fixed tracking ratio mode.

a. Normal

a. If +20V and -20V outputs are both normal with no load, a

supply might be current limiting under load. To check this adjustment see steps (l) thru (p) on page A-4. b. Proceed to -20V supply troubleshooting.

b. High, low, or zero out-

put voltage.

3 Check output of +6V supply. a. Normal

a. If the output of this supply is normal unloaded but its volt-

age falls when loaded, check the current limit adjust- ment, steps (q) thru (t) on page A-4. b. Proceed to Table A-7.

b. High, low, or zero out-

put voltage.

Table A-4. Bias and Reference Voltage Check

STEP ACTION RESPONSE NEXT ACTION

1 Check +12V bias. a. Normal (+12V ± 5%) b. Voltage high c. Output voltage lower or

a. Proceed to step (3). b. Check U3 for short. c. Check U3 for open.

+12V or -12V bias voltages.

2 Check -5V reference. a. Normal (-5V ± 2%) b. Voltage high c. Voltage low

a. Proceed to step (3). b. Check U5 for open. c. Check U5 and Q2 for short.

3 Check -12V bias. a. Normal (-12V ± 5%)

a. Proceed to +20V supply troubleshooting, Table A-5. b. Check U1 for short.

b. High voltage

c. Check U1 for open.

c. Low voltage

Table A-5. +20V Supply Troubleshooting

SYMPTOM STEP - ACTION RESPONSE PROBABLE CAUSE

High output voltage (higher than rating)

1. Attempt to turn off Q9 by shorting emitter-to-base of Q9.

a. Output voltage remains high. b. Output voltage becomes near zero

a. Q9 shorted. b. Remove short and proceed to step

volt.

(2).

2. Measure voltage at base of Q5.

a. Measured voltage is less than 0

a. Check for open CR13 or R36 and

volt. b. Measured voltage is more than 0

check for defective U6A. b. Check for defective Q5.

volt.

higher than rating · higher than rating

Page 18 of the Agilent Technologies E3630A user manual — SYMPTOM STEP - ACTION RESPONSE PROBABLE CAUSE
Page 18 · click to enlarge
Page 18

Table A-5. +20V Supply Troubleshooting (Cont’d)

Low output voltage (lower than rating)

1. Measure voltage at the base of Q5.

a. Measured voltage is less than 0

a. Check for open Q9, Q6, R26, or

volt. b. Measures voltage is more than 0

CR5. b. Proceed to step (2).

volt.

2. Eliminate current limit circuit as a source of trouble by disconnecting anode of CR11.

a. Output voltage increases. b. Output voltage remains low.

a. Check for U6B defective. b. Reconnect lead and proceed to

step (3).

3. Measure voltage at pin 3 of U6.

a. Measured voltage is near

a. Check for defective U6A.

-0.7V. b. Measured voltage is zero volt.

b. Check for CR14 or CR15 shorted. c. Check for open R39, shorted R36,

c. Measured voltage is near

or leaky or shorted C7.

+0.7V.

Table A-6. -20V Supply Troubleshooting

SYMPTOM STEP - ACTION RESPONSE PROBABLE CAUSE

The +20V supply must operate properly before troubleshooting the -20V supply.

High output voltage (more than 1% greater than +20V supply in fixed tracking ratio mode)

1. Attempt to turn off Q8 by shorting emitter-to-base of Q8.

a. Output voltage remains high. b. Output voltage becomes near zero

a. Q8 shorted. b. Remove short and proceed to

volt.

step (2).

2. Measure voltage at base of Q7.

a. Measured voltage is more than

a. Check for open CR20 or R14 and

0 volt. b. Measured voltage is less than

check for defective U8A. b. Check for defective Q7.

0 volt.

Low output voltage (more than 1% lower than +20V supply in fixed tracking ratio mode)

1. Measure voltage at the base of Q7.

a. Measured voltage is more than 0

a. Check for open Q8, Q3, and R24. b. Proceed to step (2).

volt. b. Measured voltage is less than 0 volt.

a. Check for U8B defective. b. Reconnect lead and proceed to

2. Eliminate current limit circuit as a source of trouble by disconnecting anode of CR18.

a. Output voltage increases. b. Output voltage remains low.

step (3).

a. Check for defective U8A.

3. Measure voltage at pin 3 of U8.

a. Measured voltage is near

+0.7V. b. Measured voltage is zero volt.

b. Check for shorted CR21 and

CR22. c. Check for open R15, shorted R41,

c. Measured voltage is near

or leaky or shorted C6.

Page 19 of the Agilent Technologies E3630A user manual — STEP ACTION RESPONSE PROBABLE CAUSE
Page 19 · click to enlarge
Page 19

Table A-7. +6V Supply Troubleshooting

SYMPTOM STEP - ACTION RESPONSE PROBABLE CAUSE

The +20V supply must operate properly before troubleshooting the -20V supply.

High output voltage (higher than rating)

1. Attempt to turn off Q1 by shorting emitter-to-base of Q1.

a. Output voltage remains high. b. Output voltage becomes near zero

a. Q1 shorted. b. Remove short and proceed to

volt.

step 2.

2. Measure voltage at base of Q4.

a. Measured voltage is more than

a. Check for defective Q4.

-0.6V. b. Measured voltage is less than

b. Check for open CR6, R11 and

check for defective U2A.

Low output voltage (lower than rating)

1. Measure voltage at the base of Q4.

a. Measured voltage is less than

a. Check for open Q1, Q6, R25, or

-0.6V. b. Measured voltage is more than

CR5. b. Proceed to step (2).

2. Eliminate current limit circuit as a source of trouble by disconnecting anode of CR7.

a. Output voltage increases. b. Output voltage remains low.

a. Check for U2B defective. b. Reconnect lead and proceed to

step (3).

3. Measure voltage at pin 3 of U2.

a. Measured voltage is near

a. Check for defective U2A.

-0.7V. b. Measured voltage is zero volt. c. Measured voltage is near

b. Check for CR9 and CR10 shorted. c. Check for open R29, shorted R11,

+0.7V.

or leaky or shorted C5.

Table A-8. +20V Preregulator/Control Circuit Troubleshooting

STEP ACTION RESPONSE PROBABLE CAUSE

Set output voltage at 12V+0.5V.

1 Measure pin 1 of U14. a. Measured voltage is -12V. b. Measured voltage is near

a. Proceed to step (2). b. Check for defective U14A.

+4.3V.

2 Measure pin 1 of U13. a. Measured voltage is near +1V

a. Check for defective U13 or

CR12. b. Check for open Q11 or R81.

b. Measured voltage is near 0V.

3 Measure pin 1 of U11. a. Measured voltage is near +1V b. Measured voltage is near 0V.

a. Check for defective U11. b. Check for open Q11 or R82.

Table A-9. -20V Preregulator/Control Circuit Troubleshooting

STEP ACTION RESPONSE PROBABLE CAUSE

Set output voltage at -12V-0.5V.

1 Measure pin 7 of U14. a. Measured voltage is -12V. b. Measured voltage is near

a. Proceed to step (2). b. Check for defective U14B.

+4.3V.

2 Measure pin 1 of U12. a. Measured voltage is near +1V

a. Check for defective U12 or

CR31. b. Check for open Q10 or R56

b. Measured voltage is near

+4.3V.

3 Measure pin 1 of U10. a. Measured voltage is near +1V

a. Check for defective U10 or

CR27. b. Check for open Q10 or R55.

b. Measured voltage is near 0V.

Page 20 of the Agilent Technologies E3630A user manual — ADJUSTMENT AND CALIBRATION
Page 20 · click to enlarge
Page 20

ADJUSTMENT AND CALIBRATION

c. Set the output voltage below 18V (ex, 15V), and adjust R11 on the display board until front panel VOLTS display reads exactly DVM value. Next, set the output voltage above 20V (ex, 21V) and adjust R17 on the display board until front panel VOLTS display reads exactly DVM value.

+6V Supply. To adjust the current limit circuit in the +6V sup- ply, proceed as follows: a. Check the setting of the current limit by performing steps (p) and (q) on page A-5. (Be sure to set the output voltage to 6 volts.) If reducing the load resistance permits the cur- rent to exceed 2.9 A, stop, turn R6 slightly clockwise, and repeat the test. If, instead, the current begins to fall before it reaches 2.6 A, turn R6 slightly counter clockwise and repeat the test. b. Recheck the setting and readjust R6 until the test shows that the current limit circuit begins to reduce the current when a decreasing load resistance increases it to 2.75 A ± 5%.

No other voltage output needs to be calibrated after calibrating the voltmeter of +20V output.

Panel Ammeter. To calibrate ammeter for +20V output and - 20V output, proceed as follows: a. Connect RS between +20V output terminal and COM ter- minal and connect DVM across RS. b. Push +20V meter switch and turn ±20V VOLTAGE control fully clockwise. c. Turn on the supply and adjust R5 on the display board until front panel AMPS display reads exactly DVM value divided by Rs. d. To calibrate the ammeter for +6V output, select +6V meter switch and turn +6V VOLTAGE control fully clock- wise. e. Adjust R61 on the main board until front panel AMPS dis- play reads exactly DVM value divided by Rs.

Panel Voltmeters. To calibrate voltmeter, proceed as follows: a. Connect DVM across +20V output terminal and COM ter- minal of the supply. b. Turn on the supply.

Current Limit Adjustment · Meter Calibration

Page 21 of the Agilent Technologies E3630A user manual — CODE MANUFACTURER ADDRESS
Page 21 · click to enlarge
Page 21

REPLACEABLE PARTS

Table A-10. Reference Designators

INTRODUCTION This section contains information for ordering replacement parts. Table A-12 lists parts by reference designators and pro- vides the following information:

A Assembly C Capacitor CR Diode DS Signaling Device(light) F Fuse G Pulse Generator J Jack L Inductor Q Transistor R Resistor S Switch T Transformer TP Test Point VR Zener Diode U Integrated Circuit W Wire

a. Reference designators. Refer to Table A-10. b. Agilent Technologies Part Number. c. Total quantity used in that assembly. d. Description. e. Manufacturer's supply code number. Refer to Table A-11 for manufacturer's name and address. f. Manufacturer's part number or type.

Mechanical and miscellaneous parts are not identified by ref- erence designator.

ORDERING INFORMATION To order a replacement part, address order or inquiry to your local Agilent Technologies sales office (see lists at rear of this manual for addresses). Specify the following information for each part: Model, complete serial number of the power sup- ply; Agilent Technologies part number; circuit reference des- ignator; and description.

Table A-11. Code List of Manufacturers

CODE MANUFACTURER ADDRESS

01295 Texas Instruments Inc, Semicon Comp Div. Dallas, TX

14936 General Instruments Corp, Semicon Prod Hicksville, N.Y.

27014 National Semiconductor Corporation Santa Clara, CA

28480 Agilent Technologies Palo Alto, CA

04713 Motorola Semiconductor Products Phoenix, AZ

32997 Bourns Inc. Riverside, CA

Page 22 of the Agilent Technologies E3630A user manual — Parts List
Page 22 · click to enlarge
Page 22
Agilent Technologies E3630A — Parts List diagram (page 22)
Fig. 4 · Parts List

!"# ! $ %

!"# "# # $ %

!"# # !

!

! $ %

! $ %

! !

' (

! !

! ! #

! ! !

!

)" *'# # #

)"' # #

)" *'# # #

&"

)&"*' #

Page 23 of the Agilent Technologies E3630A user manual — REFERENCE DESIGNATION
Page 23 · click to enlarge
Page 23

REFERENCE DESIGNATION

PART NUMBER

Q'TY DESCRIPTION MFR. CODE MFR. P/N

)"*' # # #

) *' )

)&"'

)

) '

)" *'# # #

)&"*' #

) '

)"' # #

)" *'# # #

)&"*'

*' +*

&*' !+*

&*' !+*

*' +*

' ,!

*'& +*

&*'& +*

&*'& +*

&*'& +*

'& +*

&'& +*

*'& +*

*'& +*

&*'& +*

*'& +*

&*'& +*

*'& +*

*'& +*

*'& +*

&*'& +*

& *'& +*

&*&'& +*

& &'& +*

& *'& +*

*'& +*

&*'& +*

" *'-& .

& &'& +*

&*&'& +*

*'& +*

*'& +*

&*'& +*

·

Page 24 of the Agilent Technologies E3630A user manual — REFERENCE DESIGNATION
Page 24 · click to enlarge
Page 24

REFERENCE DESIGNATION

PART NUMBER

Q'TY DESCRIPTION MFR. CODE MFR. P/N

&*'& +*

*'& +*

*'& +*

*&'& +*

'& +*

' ,!

'& +*

'& +*

*&'& +*

*&'& +*

*' +*

& *'& +*

" *'

" #)! &(&

" "# ! #!

" #) &(&

" # )&(& # /&

" #) &(&

" 0" ! #

" # # + ) /

" !"# ! # !

! "

!

! &

+

/!& ' + ! #

/!& ' + ! #

0 ! ! # + ,

0 ! ! !! +

0 ! ! ! +

0 ! ! !! + !

0 ! ! ! + !

0 ! ! ! +

0 ! ! +

# # !

) (( 1

Show 5 more labels / short text on this page
· ! · ! · ·
Page 25 of the Agilent Technologies E3630A user manual — Parts List
Page 25 · click to enlarge
Page 25

Agilent Technologies E3630A — Parts List diagram (page 25)
Fig. 5 · Parts List

, !! & ! !

, !! & ! !

& $ %

#!

# &

Page 26 of the Agilent Technologies E3630A user manual —
Page 26 · click to enlarge
Page 26

CERTIFICATION Agilent Technologies certifies that this product met its published specifications at time of shipment from the factory. Agilent Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Stan- dards and Technology (formerly National Bureau of Standards), to the extent allowed by that organization's calibration facility, and to the calibration facilities of other International Standards Organization members.

WARRANTY This Agilent Technologies hardware product is warranted against defects in material and workmanship for a period of three years from date of delivery. Agilent software and firmware products, which are designated by Agilent for use with a hardware product and when properly installed on that hardware product, are warranted not to fail to execute their programming instruc- tions due to defects in material and workmanship for a period of 90 days from date of delivery. During the warranty period, either Agilent or Agilent Technologies will, at its option, either repair or replace products which prove to be defective. Agilent does not warrant that operation the software, firmware, or hardware shall be uninterrupted or error free.

For warranty service, with the exception of warranty options, this product must be returned to a service facility designated by Agilent. Return to Englewood Colorado Service Center for repair in United States(1-800-258-5165). Customer shall prepay shipping charges by (and shall pay all duty and taxes) for products returned to Agilent for warranty service. Except for the products returned to Customer from another country, Agilent shall pay for return of products to Customer.

Warranty services outside the country of initial purchase are included in Agilent's product price, only if Customer pays Agilent international prices (defined as destination local currency price, or U.S. or Geneva Export price).

If Agilent is unable, within a reasonable time, to repair or replace any product to condition as warranted, the Customer shall be entitled to a refund of the purchase price upon return of the product to Agilent.

The warranty period begins on the date of delivery or on the date of installation if installed by Agilent.

LIMITATION OF WARRANTY The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by the Customer, Cus- tomer-supplied software or interfacing, unauthorized modification or misuse, operation outside of the environmental specifica- tions for the product, or improper site preparation and maintenance. TO THE EXTENT ALLOWED BY LOCAL LAW, NO OTHER WARRANTY IS EXPRESSED OR IMPLIED. AND AGILENT SPECIFICALLY DISCLAIMS THE IMPLIED WARRAN- TIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.

For consumer transactions in Australia and New Zealand: The warranty terms contained in this statement, except to the extent lawfully permitted, do not exclude, restrict or modify and are in addition to the mandatory rights applicable to the sale of this product to you.

EXCLUSIVE REMEDIES TO THE EXTENT ALLOWED BY LOCAL LAW, THE REMEDIES PROVIDED HEREIN ARE THE CUSTOMER'S SOLE AND EXCLUSIVE REMEDIES. AGILENT SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER BASED ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY.

ASSISTANCE The above statements apply only to the standard product warranty. Warranty options, extended support contacts, product maintenance agreements and customer assistance agreements are also available. Contact your nearest Agilent Technolo- gies Sales and Service office for further information on Agilent's full line of Support Programs.

Page 27 of the Agilent Technologies E3630A user manual — EN 50082-1:1992
Page 27 · click to enlarge
Page 27

DECLARATION OF CONFORMITY

according to ISO/IEC Guide 22 and EN 45014

Manufacturer’s Name: Agilent Technologies, Inc.

Manufacturer’s Address345-15, Kasan-dong, Kumchon-ku, Seoul 153-023 Korea

declares, that the products

Product NameDC Power Supply
Model NumbersE3630A
Product OptionsAll Options

conforms to the following Product Specifications:

SafetyIEC 1010-1:1990+A1:1992 / EN 61010-1:1993
EMCCISPR 11:1990 / EN 55011:1991 - Group 1 Class A 1)

EN 50082-1:1992

IEC 801-2 : 1991 - 4KV CD, 8KV AD IEC 801-3 : 1984 - 3V/m IEC 801-4 : 1988 - 1kV Power Lines

0.5kV Signal Lines

Supplementary Information: The product herewith comply with the requirements of the Low Voltage Directive 73/23/EEC and the EMC Directive 89/336/EEC and carry the “CE” mark accordingly.

1. The products was tested in a typical configuration with Agilent Technologies Test System.

Seoul, Korea November 1, 1999

European Contact for regulatory topics only: Hewlett-Packard GmbH, HQ-TRE, Herrenberger Strabe 110-140, D-71034 Böbligen (FAX: +49-7031-143143).

Quality Manager

Fix-it

Troubleshooting

Common problems, likely causes, and what to do.

ProblemPossible causeRemedy
High ripplea. Check operating setup for ground loops. b. Check main rectifiers(CR1, CR25, CR26) for open. c. Supply may be operating in current limit mode. Check current limit adjustment, steps (k) thru (r) on page A-5. p.16
Will not current limitCheck for open OR-gate diodes (CR7, CR11, CR18) or defective current limit amplifier (U2, U6, U8). p.16
Poor load and line regulationa. Check bias and reference voltages, Table A-4. b. Check main rectifiers and filters for opens. p.16
Oscillation or poor transient response timea. High frequency oscillations (above 50 kHz) can be caused by an open C11, C20, or C25. b. A defective output capacitor (C2, C3, or C4) can cause oscillations in one of many frequency ranges. c. Oscillation only in the current limiting mode can be caused by an open C12, C19, or C24. p.16
Transient voltage overshoot at turn-on or turn-offa. Overshoot only in the -20V supply can be caused by a shorted Q3. b. Overshoot in all three supply outputs can be caused by an open Q2 or a shorted Q6. p.16
Excessive heata. Check preregulator control circuit. Refer to Table A-8 and Table A-9. b. Check CR27, CR28, CR31, CR32 for short. p.16
Output Voltage clamped above 10V for ± 20V outputCheck preregulator control circuit. Refer to Table A-8 and Table A-9. p.16
High output voltage (higher than rating)a. Q9 shorted. b. Remove short and proceed to step (2). a. Check for open CR13 or R36 and check for defective U6A. b. Check for defective Q5.1. Attempt to turn off Q9 by shorting emitter-to-base of Q9. 2. Measure voltage at base of Q5. p.17
The +20V supply must operate properly before troubleshooting the -20V supply. p.18
High output voltage (more than 1% greater than +20V supply in fixed tracking ratio mode)a. Q8 shorted. b. Remove short and proceed to step (2). a. Check for open CR20 or R14 and check for defective U8A. b. Check for defective Q7.1. Attempt to turn off Q8 by shorting emitter-to-base of Q8. 2. Measure voltage at base of Q7. p.18
Low output voltage (more than 1% lower than +20V supply in fixed tracking ratio mode)a. Check for open Q8, Q3, and R24. b. Proceed to step (2). a. Check for U8B defective. b. Reconnect lead and proceed to step (3). a. Check for defective U8A. b. Check for shorted CR21 and CR22. c. Check for open R15, shorted R41, or leaky or shorted C6.1. Measure voltage at the base of Q7. 2. Eliminate current limit circuit as a source of trouble by disconnecting anode of CR18. 3. Measure voltage at pin 3 of U8. p.18
High output voltage (higher than rating)a. Q1 shorted. b. Remove short and proceed to step 2. a. Check for defective Q4. b. Check for open CR6, R11 and check for defective U2A.1. Attempt to turn off Q1 by shorting emitter-to-base of Q1. 2. Measure voltage at base of Q4. p.19
Low output voltage (lower than rating)a. Check for open Q1, Q6, R25, or CR5. b. Proceed to step (2). a. Check for U2B defective. b. Reconnect lead and proceed to step (3). a. Check for defective U2A. b. Check for CR9 and CR10 shorted. c. Check for open R29, shorted R11, or leaky or shorted C5.1. Measure voltage at the base of Q4. 2. Eliminate current limit circuit as a source of trouble by disconnecting anode of CR7. 3. Measure voltage at pin 3 of U2. p.19

Ask about the Agilent Technologies E3630A

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the Agilent Technologies E3630A

Be the first to ask — other owners or our team will answer.

Original document

Agilent Technologies E3630A manual PDF

27 pages·738 KB·Power Supply·PDF
Click any page above to open it here
Page 1 / …
100%
Loading pages…