
Agilent Technologies 85036B/E
75 Ω Type-N Calibration Kits
This manual applies directly to Agilent 85036B calibration kits with serial number prefix 3514A, and Agilent 85036E calibration kits with serial number prefix 3142A. The calibration devices in this kit are individually serialized. Record the device serial numbers in the table provided in this manual. (See “Recording the Device Serial Numbers” in Chapter 1.)

© Copyright 1995, 2001, 2002 Agilent Technologies, Inc. All rights reserved.

THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED "AS IS," AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, AGILENT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. AGILENT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR ANY INFORMATION CONTAINED HEREIN. SHOULD AGILENT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL.
Product maintenance agreements and other customer assistance agreements are available for Agilent products.
For any assistance, contact Agilent Technologies. Refer to page 5-3.
ii 85036B/E



General Information Calibration Kit Overview
The Agilent 85036B and 85036E type-N calibration kits are used to calibrate Agilent network analyzers up to 3 GHz for measurements of components with 75Ω type-N connectors.
Refer to Table 6-1 and Figure 6-1 for a complete list of kit contents and their associated part numbers.
Refer to Table 6-2 and Figure 6-2 for a complete list of kit contents and their associated part numbers.
The broadband loads are instrument-grade, 75Ω terminations that have been optimized for performance up to 3 GHz. The rugged internal structure provides for highly repeatable connections. A distributed resistive element on sapphire provides excellent stability and return loss.
The opens and shorts are built from parts that are machined to the current state-of the-art precision machining.
The short’s inner conductors have a one-piece construction, common with the shorting plane. This construction provides for extremely repeatable connections.
The female open has a separate-piece inner conductor that is made from a low-dielectric-constant plastic to minimize compensation values.
Both the opens and shorts are constructed so that the pin depth can be controlled very tightly, thereby minimizing phase errors. Some of the opens and shorts have offsets. The lengths of these offsets are designed so that the difference in phase of their reflection coefficients is approximately 180 degrees at all frequencies.

General Information Calibration Kit Overview
Like the other devices in the kit, the adapters are built to very tight tolerances to provide good broadband performance. The adapters utilize a dual-beaded connector structure to ensure stable, repeatable connections. The beads are designed to minimize return loss and are separated far enough so that interaction between the beads is minimized.
The adapters are designed so that their nominal electrical lengths are the same, which allows them to be used in calibration procedures for non-insertable devices.
The calibration kit must be selected and the calibration definitions for the devices in the kit installed in the network analyzer prior to performing a calibration. Refer to your network analyzer user’s guide for instructions on selecting the calibration kit and performing a calibration.
The calibration definitions can be:
The calibration definitions for the kit may be permanently installed in the internal memory or hard disk of the network analyzer.
If the calibration definitions for the kit are not permanently installed in the network analyzer, they must be manually entered. Refer to your network analyzer user’s guide for instructions.
Gages, torque and open-end wrenches, ESD protective devices, and various connector cleaning supplies are not included in the calibration kit but are required to ensure successful operation of the calibration kit. Refer to Table 6-3 on page 6-5 for ordering information

General Information Incoming Inspection
Verify that the shipment is complete by referring to Figure 1-1 on page 1-5.
Check for damage. The foam-lined storage case provides protection during shipping.
If the case or any device appears damaged, or if the shipment is incomplete, contact Agilent. See Table 5-1 on page 5-3. Agilent will arrange for repair or replacement of incomplete or damaged shipments without waiting for a settlement from the transportation company.
When you send the kit or device to Agilent, include a service tag (found near the end of this manual) with the following information:

General Information Incoming Inspection


General Information Recording the Device Serial Numbers
In addition to the kit serial number, the devices in this kit are individually serialized (serial numbers are labeled onto the body of each device). Record these serial numbers in Table 1-1 for the 85036B and Table 1-2 for the 85036E. Recording the serial numbers will prevent confusing the devices in this kit with similar devices in other kits.

General Information Calibration Kits Documented in this Manual
This manual applies to any 85036B or 85036E calibration kit whose serial prefix is listed on the title page. If your calibration kit has a different serial number prefix than the one listed on the title page, refer to “Calibration Kit History” below for information on how this manual applies.
This section describes calibration kits with serial number prefixes different than those listed on the title page.
85036B calibration kits with serial number prefix 3102A and below should meet the specifications published in this manual. If you wish to have your kit recertified to current specifications, contact your nearest Agilent service center. For contact information, see Table 5-1 on page 5-3.
These calibration kits had a center conductor extender with a smaller body. Therefore, a different procedure was used for connecting the female open. If you’re using one of the smaller extenders, replace steps 1 through 3 in “Connecting and Disconnecting the Two-Piece Female Open (85036B)” on page 3-9 of this manual with the steps below.
The part numbers in this manual are the recommended replacement parts for these kits (including the center conductor extender).

General Information Clarifying the Sex of a Connector
In this manual, the sex of calibration devices and adapters are referred to in terms of their connector interface. For example, a male open has a male connector.
However, during a measurement calibration, the network analyzer softkey menus label a type-N calibration device with reference to the sex of the analyzer’s test port connector—not the calibration device connector. For example, the label SHORT(F) on the analyzer’s display refers to the short that is to be connected to the female test port. This will be a male short from the calibration kit.
Conversely, connector gages are referred to in terms of the connector that it measures. For instance, a male connector gage has a female connector on the gage so that it can measure male devices.
The best techniques for maintaining the integrity of the devices in this kit include:
All of the above are described in Chapter 3 , “Use, Maintenance, and Care of the Devices.” Failure to detect and remove dirt or metallic particles on a mating plane surface can degrade repeatability and accuracy and can damage any connector mated to it. Improper connections, resulting from pin depth values being out of the limits (see Table 2-3 on page 2-3), or from bad connections, can also damage these devices.


+18 °C to +28 °C (+64 °F to +82 °F)
±1 °C of measurement calibration temperature
Error-corrected temperature rangeb
Storage temperature −40 °C to +75 °C (−40 °F to +167 °F)
Operation < 4,500 meters (≈15,000 feet)
Storage < 15,000 meters (≈50,000 feet)
Relative humidity Always non-condensing
Operation 0 to 80% (26 °C maximum dry bulb)
Storage 0 to 95%
a. The temperature range over which the calibration standards maintain conformance to their
specifications.
b. The allowable network analyzer ambient temperature drift during measurement calibration
and during measurements when the network analyzer error correction is turned on. Also, the range over which the network analyzer maintains its specified performance while correction is turned on.
Changes in temperature can affect electrical characteristics. Therefore, the operating temperature is a critical factor in performance. During a measurement calibration, the temperature of the calibration devices must be stable and within the range specified in Table 2-1.

The mechanical specifications in Table 2-2 apply to the devices in the 85036B and the 85036E 75 Ω type-N calibration kits.
Type-N Male Open Inside diameter of outer conductor: 6.985 to 7.010
Type-N Male Short Inside diameter of outer conductor: 7 ±0.015 mm Distance from reference plane to shorting plane: 5.29 ±0.013 mm
Supplemental characteristics are values which are typically met by a majority of the calibration kit devices tested at Agilent. These supplemental characteristics are intended to provide information useful in calibration kit applications by giving typical, but non-warranted performance parameters. Table 2-3 lists the typical characteristics of the devices in this kit.
Type-N Male Short Diameter of male pin: 0.897 ±0.013 mm
All Type-N Male Connectors Pin depth: 0.207 to 0.210 inch
All Type-N Female Connectors Pin depth: 0.204 to 0.207 inch
Specifications Mechanical Specifications · Device Specification · Device Characteristic

The electrical specifications in Table 2-4 apply to the devices in your calibration kit when connected with an Agilent precision interface.
Device Specification Frequency (GHz)
Return loss ≥ 46 dB (ρ ≤ 0.00501) DC to ≤ 2
Return loss ≥ 40 dB (ρ ≤ 0.01000) > 2 to ≤ 3
a. 23° +5 °C; typical resistance change: ±300 ppm/°C
Agilent Technologies certifies that this product met its published specifications at the time of shipment from the factory. Agilent further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology (NIST) to the extent allowed by the institute’s calibration facility, and to the calibration facilities of other International Standards Organization members. See “How Agilent Verifies the Devices in This Kit” on page 4-2 for more information.
Specifications Electrical Specifications · Loadsa

Devices

Use, Maintenance, and Care of the Devices Electrostatic Discharge
Protection against ESD (electrostatic discharge) is essential while connecting, inspecting, or cleaning connectors attached to a static-sensitive circuit (such as those found in test sets).
Static electricity can build up on your body and can easily damage sensitive internal circuit elements when discharged. Static discharges too small to be felt can cause permanent damage. Devices such as calibration components and devices under test (DUTs), can also carry an electrostatic charge. To prevent damage to the test set, components, and devices:
Refer to Chapter 6 , “Replaceable Parts,” for part numbers and instructions for ordering ESD protective devices.


Use, Maintenance, and Care of the Devices
Visual inspection and, if necessary, cleaning should be done every time a connection is made. Metal particles from the connector threads may fall into the connector when it is disconnected. One connection made with a dirty or damaged connector can damage both connectors beyond repair.
In some cases, magnification is necessary to see damage on a connector; a magnifying device with a magnification of ≥10× is recommended. However, not all defects that are visible only under magnification will affect the electrical performance of the connector. Use the following guidelines when evaluating the integrity of a connector.
Examine the connectors first for obvious defects and damage: badly worn plating on the connector interface, deformed threads, or bent, broken, or misaligned center conductors. Connector nuts should move smoothly and be free of burrs, loose metal particles, and rough spots.
Connector wear is caused by connecting and disconnecting the devices. The more use a connector gets, the faster it wears and degrades. The wear is greatly accelerated when connectors are not kept clean, or are connected incorrectly.
Connector wear eventually degrades performance of the device. Calibration devices should have a long life if their use is on the order of a few times per week. Replace devices with worn connectors.
The test port connectors on the network analyzer test set may have many connections each day, and are therefore more subject to wear. It is recommended that an adapter be used as a test port saver to minimize the wear on the test set’s test port connectors.
Flat contact between the connectors at all points on their mating plane surfaces is required for a good connection. Look especially for deep scratches or dents, and for dirt and metal particles on the connector mating plane surfaces. Also look for signs of damage due to excessive or uneven wear or misalignment.
Light burnishing of the mating plane surfaces is normal, and is evident as light scratches or shallow circular marks distributed more or less uniformly over the mating plane surface. Other small defects and cosmetic imperfections are also normal. None of these affect electrical or mechanical performance.
If a connector shows deep scratches or dents, particles clinging to the mating plane surfaces, or uneven wear, clean and inspect it again. Devices with damaged connectors should be discarded. Determine the cause of damage before connecting a new, undamaged connector in the same configuration.
Visual Inspection

Use, Maintenance, and Care of the Devices Cleaning Connectors
Precision slotless female connectors are used to improve accuracy. The slotless contacts are not affected by the slight variations in male contact pin diameter. However, it is still advisable to inspect them regularly for damage.
Clean connectors are essential for ensuring the integrity of RF and microwave coaxial connections.
Use compressed air (or nitrogen) to loosen particles on the connector mating plane surfaces.
You can use any source of clean, dry, low-pressure compressed air or nitrogen that has an effective oil-vapor filter and liquid condensation trap placed just before the outlet hose.
Ground the hose nozzle to prevent electrostatic discharge, and set the air pressure to less than 414 kPa (60 psi) to control the velocity of the air stream. High-velocity streams of compressed air can cause electrostatic effects when directed into a connector. These electrostatic effects can damage the device. Refer to “Electrostatic Discharge” earlier in this chapter for additional information.
Use isopropyl alcohol with adequate ventilation and avoid contact with eyes, skin, and clothing. It causes skin irritation, may cause eye damage, and is harmful if swallowed or inhaled. It may be harmful if absorbed through the skin. Wash thoroughly after handling.
In case of spill, soak up with sand or earth. Flush spill area with water.
Dispose of isopropyl alcohol in accordance with all applicable federal, state, and local environmental regulations.

Use, Maintenance, and Care of the Devices
Use a lint-free swab or cleaning cloth moistened with isopropyl alcohol to remove any dirt or stubborn contaminants on a connector that cannot be removed with compressed air or nitrogen. Refer to Table 6-3 on page 6-5 for part numbers for isopropyl alcohol and cleaning swabs.
a. Apply a small amount of isopropyl alcohol to a lint-free cleaning swab.
b. Clean the connector threads.
c. Let the alcohol evaporate, then blow the threads dry with a gentle stream of clean,
low-pressure compressed air or nitrogen. Always completely dry a connector before you reassemble or use it.
a. Apply a small amount of isopropyl alcohol to a lint-free cleaning swab.
b. Clean the center and outer conductor mating plane surfaces. When cleaning a female
connector, avoid snagging the swab on the center conductor contact fingers by using short strokes.
c. Let the alcohol evaporate, then blow the connector dry with a gentle stream of clean,
low-pressure compressed air or nitrogen. Always completely dry a connector before you reassemble or use it.
Inspect the connector again to make sure that no particles or residue are present.
Cleaning Connectors

Use, Maintenance, and Care of the Devices Connections
Good connections require a skilled operator. The most common cause of measurement error is bad connections. The following procedures illustrate how to make good connections.
Do not twist or screw the connectors together. As the center conductors mate, there is usually a slight resistance.
A connection in which the outer conductors make gentle contact at all points on both mating surfaces is sufficient. Very light finger pressure is enough to accomplish this.

Use, Maintenance, and Care of the Devices
Connector Type Torque Setting Torque Tolerance
Type-N 135 N-cm (12 in-lb) ±13.5 N-cm (±1.2 in-lb)
Using a torque wrench guarantees that the connection is not too tight, preventing possible connector damage. It also guarantees that all connections are equally tight each time.
Connections

Use, Maintenance, and Care of the Devices Connections
Do not hold the wrench so tightly that you push the handle straight down along its length rather than pivoting it, otherwise you apply an unknown amount of torque.
Do not pivot the wrench handle on your thumb or other fingers, otherwise you apply an unknown amount of torque to the connection when the wrench reaches its break point.
Do not twist the head of the wrench relative to the outer conductor mating plane. If you do, you apply more than the recommended torque.

Use, Maintenance, and Care of the Devices
The female open standard in the 85036B calibration kit is composed of two parts: the open body (outer conductor) and the center conductor extender. Refer to Figure 3-4.
To connect the female open:
To disconnect the female open:
To avoid lateral (bending) force on the connector mating plane surfaces, always support the devices and connections.
Connections

Use, Maintenance, and Care of the Devices Handling and Storage


The performance of your calibration kit can only be verified by returning the kit to Agilent Technologies for recertification. The equipment required to verify the specifications of the devices in the kit has been specially manufactured and is not commercially available.
Agilent verifies the specifications of these devices as follows:
These two steps establish a traceable link to NIST for Agilent to the extent allowed by the institute’s calibration facility. The specifications data provided for the devices in this kit is traceable to NIST through Agilent Technologies.
Performance Verification Introduction

The following will be provided with a recertified kit:
Agilent Technologies offers a Standard calibration for the recertification of this kit. For more information, contact Agilent Technologies. See Table 5-1 on page 5-3.
The 75Ω loads in the calibration kit are specified for use up to 3 GHz. For many applications, the performance above 2 GHz is not utilized.
For a standard recertification, the devices in the calibration kit are tested and calibrated up to 3 GHz. However, a limited recertification can be requested. For this limited recertification, the devices are tested and calibrated up to 2 GHz.
To request a limited recertification, make sure the following is clearly written on the order: Limited Calibration DC–2 GHz.
All loads that receive a limited calibration are supplied with a limited calibration label applied to the device.
The suggested initial interval for recertification is 12 months or sooner. The actual need for recertification depends on the use of the kit. After reviewing the results of the initial recertification, you may establish a different recertification interval that reflects the usage and wear of the kit.
Contact Agilent Technologies for information on where to send your kit for recertification. See Table 5-1 on page 5-3.
When you return the kit, complete and attach a service tag. Refer to “Returning a Kit or Device to Agilent” on page 5-3 for details.
Performance Verification · Recertification

Performance Verification Recertification


If you suspect a bad calibration, or if your network analyzer does not pass performance verification, follow the steps in Figure 5-1.

Troubleshooting Troubleshooting Process

Troubleshooting Returning a Kit or Device to Agilent
If your kit or device requires service, contact the Agilent Technologies office nearest you for information on where to send it. See Table 5-1. Include a service tag (located near the end of this manual) on which you provide the following information:
United States (tel) 1 800 452 4844
Latin America (tel) (305) 269 7500 (fax) (305) 269 7599
Canada (tel) 1 877 894 4414 (fax) (905) 282-6495
Europe (tel) (+31) 20 547 2323 (fax) (+31) 20 547 2390
New Zealand (tel) 0 800 738 378 (fax) (+64) 4 495 8950
Japan (tel) (+81) 426 56 7832 (fax) (+81) 426 56 7840
Australia (tel) 1 800 629 485 (fax) (+61) 3 9210 5947
Singapore (tel) 1 800 375 8100 (fax) (65) 836 0252
Malaysia (tel) 1 800 828 848 (fax) 1 800 801 664
Philippines (tel) (632) 8426802 (tel) (PLDT subscriber only): 1 800 16510170 (fax) (632) 8426809 (fax) (PLDT subscriber only): 1 800 16510288
Hong Kong (tel) 800 930 871 (fax) (852) 2506 9233
Taiwan (tel) 0800-047-866 (fax) (886) 2 25456723
People’s Republic of China (tel) (preferred): 800-810-0189 (tel) (alternate): 10800-650-0021 (fax) 10800-650-0121
India (tel) 1-600-11-2929 (fax) 000-800-650-1101

Troubleshooting Returning a Kit or Device to Agilent


Table 6-1 lists the replacement part numbers for items included in the 85036B calibration kit and Figure 6-1 illustrates each of these items.
Table 6-2 lists the replacement part numbers for items included in the 85036E calibration kit and Figure 6-2 illustrates each of these items.
Table 6-3 lists the replacement part numbers for items recommended or required for successful operation but not included in the calibration kit.
To order a listed part, note the description, the part number, and the quantity desired. Telephone or send your order to Agilent Technologies. See Table 5-1 on page 5-3.
Description Qty Per Kit Agilent Part Number
Calibration Devices (75Ω Type-N)
Male broadband load 1 00909-60019
Female broadband load 1 00909-60020
Male short 1 85036-60012
Female short 1 85036-60011
Male open 1 85032-60007
Female open body 1 85032-20001
Female open center conductor extender 1 85036-60019
Type-N-male to male 1 85036-60013
Type-N-female to female 1 85036-60014
Type-N-male to female 1 85036-60015
Calibration Kit Storage Case
Foam pad set 1 85036-80002
User’s and service guide 1 85036-90016
Male protective end caps as required 1401-0214
Female protective end caps as required 1401-0225
Tube package (for center conductor extender) 1 1540-1027


Replaceable Parts · Introduction

Description Qty Per Kit Agilent Part Number
Calibration Devices (75Ω Type-N)
Male broadband load 1 00909-60019
Male combination open/short 1 85036-60016
Calibration Kit Storage Case
Foam pad set 1 85023-80005
User’s and service guide 1 85036-90016
Male protective end cap as required


Table 6-3 Replaceable Parts—Items Not Included in the 85036B and 85036E Calibration Kits
Description Qty Agilent Part Number
Connector Gages (75Ω Type-N)
Connector gage set 1 85036B K01
3/4 in, 135 N-cm (12 in-lb) torque wrench 1 8710-1766
1/2 in and 9/16 in open-end wrench 1 8710-1770
3/4 in open-end wrench 1 8720-0011
Grounding wrist strap 1 9300-1367
5 ft grounding cord for wrist strap 1 9300-0980
2 ft by 4 ft conductive table mat with 15 ft grounding wire 1 9300-0797
ESD heel strap 1 9300-1308
Isopropyl alcohol 30 ml 8500-5344
Foam tipped cleaning swabs 100 9301-1243
Connector care - quick reference card 1 08510-90360

Replaceable Parts Introduction


Standard Definitions Standard Class Assignments
Class assignment organizes calibration standards into a format compatible with the error models used in the measurement calibration. A class or group of classes corresponds to the systematic errors to be removed from the measured network analyzer response. Table A-1 lists the classes of the devices in this calibration kit.
Class A B C D E F G Standard Class Label
S11A 2 8 Opens
S11B 1 7 Shorts
S11C 3 Load
S22A 2 8 Opens
S22B 1 7 Shorts
S22C 3 Load
Forward Transmission 4 Thru
Reverse Transmission 4 Thru
Forward Match 4 Thru
Reverse Match 4 Thru
Response 1 7 2 8 4 Response
Response and Isolation 1 7 2 8 4 Response & Isolation

Standard Definitions Standard Class Assignments
The standard class assignments listed in Table A-1 may be changed to meet your specific requirements. Table A-2 is provided to record the modified standard class assignments.
Class A B C D E F G Standard Class Label
S 11 B
S 11 C
S 22 A
S 22 B
S 22 C
85036B/E A-3

Standard Definitions Nominal Standard Definitions
Standard definitions provide the constants needed to mathematically model the electrical characteristics (delay, attenuation, and impedance) of each calibration standard. The nominal values of these constants are theoretically derived from the physical dimensions and material of each calibration standard, or from actual measured response. These values are used to determine the measurement uncertainties of the network analyzer. The standard definitions in Table A-3 list typical calibration kit parameters used to specify the mathematical model of each device. This information must be loaded into the network analyzer to perform valid calibrations. Refer to your network analyzer’s user’s guide for instructions on loading calibration constants.
This kit contains only 75 ohm devices. Ensure the system impedance (Z0) is set to 75 ohms. Refer to your network analyzer’s user’s guide for instructions on setting system impedance.

Standard Definitions Nominal Standard Definitions
a = 75 Ω Calibration Kit Label: N 75 Ω
C3 ×10−45 F/Hz3
C1 ×10−27 F/Hz
C2 ×10−36 F/Hz2
Fixed or Slidingc
Loss (Ω/s)
1 Short 0 75 1.13G 0 999 Coax Short (m)d
2 Open 63.5 84 56 0 0 75 1.13G 0 999 Coax Open (m)d
3 Load Fixed 0 75 1.13G 0 999 Coax Broad- band
0 75 1.13G 0 999 Coax Thru
7 Short 17.544 75 1.13G 0 999 Coax Short (f)d
8 Open 41e 40 5 0 17.544 75 1.13G 0 999 Coax Open (f)d
a. Ensure system impedance (Z0) of network analyzer is set to this value. b. Open, short, load, delay/thru, or arbitrary impedance. c. Load or arbitrary impedance only. d. Standard labels that specify sex, (m) or (f), refer to the sex of the analyzer’s test port connector. e. If your instrument is an Agilent 8752B or 8711 Option 1EC, and you are calibrating at the test port, this number is 33. For all other instruments and applications, including calibrating at a point other than the test port (such as the end of a cable), 41 is the correct number. This is because the 8752B and 8711 Option 1EC have special female test ports with longer, more rugged fingers.

Standard Definitions Nominal Standard Definitions
The standard definitions listed in Table A-3 may be changed to meet your specific requirements. Table A-4 is provided to record the modified standard definitions.
C3 ×10−45 F/Hz3
C1 ×10−27 F/Hz
C2 ×10−36 F/Hz2
Fixed or Slidingc
Loss (GΩ/s)
a. Ensure system Z0 of network analyzer is set to this value.
b. Open, short, load, delay/thru, or arbitrary impedance.
c. Load or arbitrary impedance only.
d. Standard labels that specify sex, (m) or (f), refer to the sex of the analyzer’s test port connector.

when making connections, 3-6 environmental requirements, 2-2 equipment
limited calibration, 4-3 limited calibration
requesting, 4-3 loads
but not supplied, 1-3, 6-2, 6-5 supplied, 1-2, 6-2, 6-4 ESD, 3-2
M maintenance, 3-2
precautions, 3-2, 3-4 supplies
preventive, 1-8 maintenance of devices, 1-8 making connections, 3-6 manual
part numbers, 6-5 when making connections, 3-6
F female open, 3-9
part number, 6-5 mating plane
connecting, 3-9 disconnecting, 3-9 frequency
connector, 3-5 mating plane inspection, 3-3 mating plane surfaces, 3-5 mechanical specifications, 2-3
specifications, 2-4 frequency range, 2-4, 4-3
N nitrogen, 3-4 nominal standard definitions, A-4 numbers
master, 6-5 gage master
part numbers, 6-5 gages, 1-3
sales and service, 5-3 open-end wrench, 1-3, 3-9
H handling, 3-10 history, calibration kit, 1-7 humidity, 2-2
part number, 6-5 opens, 1-2 opens and shorts, 1-2 ordering parts, 6-2
P part numbers, 6-2 parts
system, A-4 incoming inspection, 1-4 inside diameter of outer conductor, specifications, 2-3 inspection
ordering, 6-2 replaceable, 6-2 required but not supplied, 6-2 performance verification, 4-2
damage, 3-3 defects, 3-3 incoming, 1-4 mating plane, 3-3 visual, 3-3 isopropyl alcohol
fail, 5-2 permanently stored calibration definitions, 1-3 pin depth
typical values, 2-3 precision slotless connectors, 3-4 preventive maintenance, 1-8
as cleaning solvent, 3-4
R recertification, 4-3
contents, 1-2 overview, 1-2 return to Agilent, 5-3 serial number, 1-6
how often?, 4-3 interval, 4-3 limited, 4-3 what’s included, 4-3 where to send your kit, 4-3 regulations
environmental, 3-5 replaceable parts, 6-2
part number, 6-5

device, 2-2 error-corrected temperature range, 2-2 measurement, 2-2 operating range, 2-2 verification and measurement, 2-2 test data, 4-2 threads
environmental, 2-2 return loss
specifications, 2-4 return, kit or device, 5-3
connector, 3-4 torque wrench, 1-3, 3-6
and service offices, 5-3 sales and service offices, 5-3 serial numbers, 1-6
part number, 6-5 specifications, 3-6 traceability, 4-2 troubleshooting, 5-2 two-piece female open, 3-9
devices, 1-6 recording, 1-6 service, 5-2
and sales offices, 5-3 service tag, 1-4, 4-3, 5-3 sex
connector, 1-8 shipment
performance, 4-2 temperature, 2-2 visual inspection, 3-3
damage, 1-4 verifying complete, 1-4 shorts, 1-2 specifications, 2-2
W warranty, documentation, ii wrench
altitude, 2-2 certification, 2-4 distance from ref plane to shorting plane, 2-3 electrical, 2-4 environmental, 2-2 frequency, 2-4 humidity, 2-2 inside diameter of outer conductor, 2-3 mechanical, 2-3 return loss, 2-4 temperature, 2-2 verifying, 4-2 standard class assignments, A-2
open-end, 1-3, 3-9
part number, 6-5 undoing connections, 3-9 torque, 1-3
part number, 6-5 wrenches
part numbers, 6-5
blank form, A-3 standard definitions, A-4, A-6
blank form, A-6 nominal, A-4 standards
calibration, 2-4 NIST, 2-4, 4-3 static discharge, 3-2 sticker
calibration, 4-3 storage, 3-10
altitude, 2-2 humidity, 2-2 temperature, 2-2 supplies
cleaning, 1-3 system impedance
service, 1-4, 4-3, 5-3 temperature
Owners and our team usually reply within a day — and your question helps the next person too.
Be the first to ask — other owners or our team will answer.