EN
Agilent Technologies logo
Official user manual

Agilent Technologies E

49 pages·809 KB·Work Light·PDF·ENDEFRITNLJA
Agilent Technologies E
Table of Contents 50 sections Title Page · Contents · 1 General Information · Calibration Kit Overview · … ShowHide
Page 1 of the Agilent Technologies E user manual — Manufacturing Part Number: 85036-90016
Page 1 · click to enlarge
Page 1

User’s and Service Guide

Agilent Technologies 85036B/E

75 Ω Type-N Calibration Kits

This manual applies directly to Agilent 85036B calibration kits with serial number prefix 3514A, and Agilent 85036E calibration kits with serial number prefix 3142A. The calibration devices in this kit are individually serialized. Record the device serial numbers in the table provided in this manual. (See “Recording the Device Serial Numbers” in Chapter 1.)

Agilent Technologies E — User’s and Service Guide diagram (page 1)
Fig. 1 · User’s and Service Guide

Manufacturing Part Number: 85036-90016

Printed in USA Print DateJune 2002 Supersedes: October 2001

© Copyright 1995, 2001, 2002 Agilent Technologies, Inc. All rights reserved.

Page 2 of the Agilent Technologies E user manual — Assistance
Page 2 · click to enlarge
Page 2

Documentation Warranty

THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED "AS IS," AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, AGILENT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. AGILENT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR ANY INFORMATION CONTAINED HEREIN. SHOULD AGILENT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL.

Assistance

Product maintenance agreements and other customer assistance agreements are available for Agilent products.

For any assistance, contact Agilent Technologies. Refer to page 5-3.

ii 85036B/E

Page 5 of the Agilent Technologies E user manual — 1 General Information
Page 5 · click to enlarge
Page 5

1 General Information

Page 6 of the Agilent Technologies E user manual — Opens and Shorts
Page 6 · click to enlarge
Page 6

General Information Calibration Kit Overview

Calibration Kit Overview

The Agilent 85036B and 85036E type-N calibration kits are used to calibrate Agilent network analyzers up to 3 GHz for measurements of components with 75Ω type-N connectors.

Kit Contents

The 85036B calibration kit contains the following:

  • one male and one female 75 Ω type-N open termination
  • one male and one female 75 Ω type-N short termination
  • one male and one female 75 Ω type-N load
  • three 75 Ω type-N to 75 Ω type-N adapters

Refer to Table 6-1 and Figure 6-1 for a complete list of kit contents and their associated part numbers.

The 85036E calibration kit contains the following:

  • one male combination open/short termination
  • one male 75 Ω type-N load

Refer to Table 6-2 and Figure 6-2 for a complete list of kit contents and their associated part numbers.

Broadband Loads

The broadband loads are instrument-grade, 75Ω terminations that have been optimized for performance up to 3 GHz. The rugged internal structure provides for highly repeatable connections. A distributed resistive element on sapphire provides excellent stability and return loss.

Opens and Shorts

The opens and shorts are built from parts that are machined to the current state-of the-art precision machining.

The short’s inner conductors have a one-piece construction, common with the shorting plane. This construction provides for extremely repeatable connections.

The female open has a separate-piece inner conductor that is made from a low-dielectric-constant plastic to minimize compensation values.

Both the opens and shorts are constructed so that the pin depth can be controlled very tightly, thereby minimizing phase errors. Some of the opens and shorts have offsets. The lengths of these offsets are designed so that the difference in phase of their reflection coefficients is approximately 180 degrees at all frequencies.

Page 7 of the Agilent Technologies E user manual — Equipment Required but Not Supplied
Page 7 · click to enlarge
Page 7

General Information Calibration Kit Overview

Adapters

Like the other devices in the kit, the adapters are built to very tight tolerances to provide good broadband performance. The adapters utilize a dual-beaded connector structure to ensure stable, repeatable connections. The beads are designed to minimize return loss and are separated far enough so that interaction between the beads is minimized.

The adapters are designed so that their nominal electrical lengths are the same, which allows them to be used in calibration procedures for non-insertable devices.

Calibration Definitions

The calibration kit must be selected and the calibration definitions for the devices in the kit installed in the network analyzer prior to performing a calibration. Refer to your network analyzer user’s guide for instructions on selecting the calibration kit and performing a calibration.

The calibration definitions can be:

  • resident within the analyzer
  • entered from the front panel

Installation of the Calibration Definitions

The calibration definitions for the kit may be permanently installed in the internal memory or hard disk of the network analyzer.

If the calibration definitions for the kit are not permanently installed in the network analyzer, they must be manually entered. Refer to your network analyzer user’s guide for instructions.

Equipment Required but Not Supplied

Gages, torque and open-end wrenches, ESD protective devices, and various connector cleaning supplies are not included in the calibration kit but are required to ensure successful operation of the calibration kit. Refer to Table 6-3 on page 6-5 for ordering information

Page 8 of the Agilent Technologies E user manual — Incoming Inspection
Page 8 · click to enlarge
Page 8

General Information Incoming Inspection

Incoming Inspection

Verify that the shipment is complete by referring to Figure 1-1 on page 1-5.

Check for damage. The foam-lined storage case provides protection during shipping.

If the case or any device appears damaged, or if the shipment is incomplete, contact Agilent. See Table 5-1 on page 5-3. Agilent will arrange for repair or replacement of incomplete or damaged shipments without waiting for a settlement from the transportation company.

When you send the kit or device to Agilent, include a service tag (found near the end of this manual) with the following information:

  • your company name and address
  • the name of a technical contact person within your company, and the person's complete phone number
  • the model number and serial number of the kit
  • the part number and serial number of the device
  • the type of service required
  • a detailed description of the problem
Page 9 of the Agilent Technologies E user manual — Figure 1-1 Calibration Kit Contents
Page 9 · click to enlarge
Page 9

General Information Incoming Inspection

Figure 1-1 Calibration Kit Contents

Agilent Technologies E — Figure 1-1 Calibration Kit Contents diagram (page 9)
Fig. 2 · Figure 1-1 Calibration Kit Contents
Page 10 of the Agilent Technologies E user manual — Table 1-2 Serial Number Record for 85036E
Page 10 · click to enlarge
Page 10

General Information Recording the Device Serial Numbers

Recording the Device Serial Numbers

In addition to the kit serial number, the devices in this kit are individually serialized (serial numbers are labeled onto the body of each device). Record these serial numbers in Table 1-1 for the 85036B and Table 1-2 for the 85036E. Recording the serial numbers will prevent confusing the devices in this kit with similar devices in other kits.

Table 1-1 Serial Number Record for 85036B

DeviceSerial Number Calibration kit Male broadband load Female broadband load Male open Female open body Male short Female short Type-N male to male adapter Type-Nfemale to female adapter Type-Nmale to female adapter

Table 1-2 Serial Number Record for 85036E

DeviceSerial Number Calibration kit Male broadband load Male combination open/short
Page 11 of the Agilent Technologies E user manual — Agilent 85036B kits with serial prefix 3103A and below
Page 11 · click to enlarge
Page 11

General Information Calibration Kits Documented in this Manual

Calibration Kits Documented in this Manual

This manual applies to any 85036B or 85036E calibration kit whose serial prefix is listed on the title page. If your calibration kit has a different serial number prefix than the one listed on the title page, refer to “Calibration Kit History” below for information on how this manual applies.

Calibration Kit History

This section describes calibration kits with serial number prefixes different than those listed on the title page.

Agilent 85036B kits with serial prefix 3102A and below

  • 75 ohm load specification: return loss ≥ 45 db, dc to 2 GHz
  • all other kit components unspecified

85036B calibration kits with serial number prefix 3102A and below should meet the specifications published in this manual. If you wish to have your kit recertified to current specifications, contact your nearest Agilent service center. For contact information, see Table 5-1 on page 5-3.

Agilent 85036B kits with serial prefix 3103A and below

These calibration kits had a center conductor extender with a smaller body. Therefore, a different procedure was used for connecting the female open. If you’re using one of the smaller extenders, replace steps 1 through 3 in “Connecting and Disconnecting the Two-Piece Female Open (85036B)” on page 3-9 of this manual with the steps below.

1. Mate the center conductor extender to the male pin of the test port.
2. Connect the open body to the test port by carefully guiding it over the extender.
3. To disconnect the open, carefully remove the open body first.

The part numbers in this manual are the recommended replacement parts for these kits (including the center conductor extender).

Page 12 of the Agilent Technologies E user manual — Preventive Maintenance
Page 12 · click to enlarge
Page 12

General Information Clarifying the Sex of a Connector

Clarifying the Sex of a Connector

In this manual, the sex of calibration devices and adapters are referred to in terms of their connector interface. For example, a male open has a male connector.

However, during a measurement calibration, the network analyzer softkey menus label a type-N calibration device with reference to the sex of the analyzer’s test port connector—not the calibration device connector. For example, the label SHORT(F) on the analyzer’s display refers to the short that is to be connected to the female test port. This will be a male short from the calibration kit.

Conversely, connector gages are referred to in terms of the connector that it measures. For instance, a male connector gage has a female connector on the gage so that it can measure male devices.

Preventive Maintenance

The best techniques for maintaining the integrity of the devices in this kit include:

  • routine visual inspection
  • cleaning
  • proper gaging
  • proper connection techniques

All of the above are described in Chapter 3 , “Use, Maintenance, and Care of the Devices.” Failure to detect and remove dirt or metallic particles on a mating plane surface can degrade repeatability and accuracy and can damage any connector mated to it. Improper connections, resulting from pin depth values being out of the limits (see Table 2-3 on page 2-3), or from bad connections, can also damage these devices.

Page 13 of the Agilent Technologies E user manual — 2 Specifications
Page 13 · click to enlarge
Page 13

2 Specifications

Page 14 of the Agilent Technologies E user manual — Temperature—What to Watch Out For
Page 14 · click to enlarge
Page 14

Environmental Requirements

Table 2-1 Environmental Requirements

+18 °C to +28 °C (+64 °F to +82 °F)

±1 °C of measurement calibration temperature

Error-corrected temperature rangeb

Storage temperature −40 °C to +75 °C (−40 °F to +167 °F)

Operation < 4,500 meters (≈15,000 feet)

Storage < 15,000 meters (≈50,000 feet)

Relative humidity Always non-condensing

Operation 0 to 80% (26 °C maximum dry bulb)

Storage 0 to 95%

a. The temperature range over which the calibration standards maintain conformance to their

specifications.

b. The allowable network analyzer ambient temperature drift during measurement calibration

and during measurements when the network analyzer error correction is turned on. Also, the range over which the network analyzer maintains its specified performance while correction is turned on.

Temperature—What to Watch Out For

Changes in temperature can affect electrical characteristics. Therefore, the operating temperature is a critical factor in performance. During a measurement calibration, the temperature of the calibration devices must be stable and within the range specified in Table 2-1.

Show 4 more labels / short text on this page
Specifications Environmental Requirements · Parameter Limits · Operating temperaturea · Altitude
Page 15 of the Agilent Technologies E user manual — Table 2-3 Mechanical Characteristics
Page 15 · click to enlarge
Page 15

Mechanical Specifications

The mechanical specifications in Table 2-2 apply to the devices in the 85036B and the 85036E 75 Ω type-N calibration kits.

Table 2-2 Mechanical Specifications

Type-N Male Open Inside diameter of outer conductor: 6.985 to 7.010

Type-N Male Short Inside diameter of outer conductor: 7 ±0.015 mm Distance from reference plane to shorting plane: 5.29 ±0.013 mm

Supplemental Mechanical Characteristics

Supplemental characteristics are values which are typically met by a majority of the calibration kit devices tested at Agilent. These supplemental characteristics are intended to provide information useful in calibration kit applications by giving typical, but non-warranted performance parameters. Table 2-3 lists the typical characteristics of the devices in this kit.

Table 2-3 Mechanical Characteristics

Type-N Male Short Diameter of male pin: 0.897 ±0.013 mm

All Type-N Male Connectors Pin depth: 0.207 to 0.210 inch

All Type-N Female Connectors Pin depth: 0.204 to 0.207 inch

Specifications Mechanical Specifications · Device Specification · Device Characteristic

Page 16 of the Agilent Technologies E user manual — Certification
Page 16 · click to enlarge
Page 16

Electrical Specifications

The electrical specifications in Table 2-4 apply to the devices in your calibration kit when connected with an Agilent precision interface.

Table 2-4 Electrical Specifications

Device Specification Frequency (GHz)

Return loss ≥ 46 dB (ρ ≤ 0.00501) DC to ≤ 2

Return loss ≥ 40 dB (ρ ≤ 0.01000) > 2 to ≤ 3

a. 23° +5 °C; typical resistance change: ±300 ppm/°C

Certification

Agilent Technologies certifies that this product met its published specifications at the time of shipment from the factory. Agilent further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology (NIST) to the extent allowed by the institute’s calibration facility, and to the calibration facilities of other International Standards Organization members. See “How Agilent Verifies the Devices in This Kit” on page 4-2 for more information.

Specifications Electrical Specifications · Loadsa

Page 17 of the Agilent Technologies E user manual — 3 Use, Maintenance, and Care of the
Page 17 · click to enlarge
Page 17

3 Use, Maintenance, and Care of the

Devices

Page 18 of the Agilent Technologies E user manual — Figure 3-1 ESD Protection Setup
Page 18 · click to enlarge
Page 18

Use, Maintenance, and Care of the Devices Electrostatic Discharge

Electrostatic Discharge

Protection against ESD (electrostatic discharge) is essential while connecting, inspecting, or cleaning connectors attached to a static-sensitive circuit (such as those found in test sets).

Static electricity can build up on your body and can easily damage sensitive internal circuit elements when discharged. Static discharges too small to be felt can cause permanent damage. Devices such as calibration components and devices under test (DUTs), can also carry an electrostatic charge. To prevent damage to the test set, components, and devices:

  • always wear a grounded wrist strap having a 1 M Ω resistor in series with it when handling components and devices or when making connections to the test set.
  • always use a grounded, conductive table mat while making connections.
  • always wear a heel strap when working in an area with a conductive floor. If you are uncertain about the conductivity of your floor, wear a heel strap.
  • always ground yourself before you clean, inspect, or make a connection to a static-sensitive device or test port. You can, for example, grasp the grounded outer shell of the test port or cable connector briefly.
  • always ground the center conductor of a test cable before making a connection to the analyzer test port or other static-sensitive device. This can be done as follows:
1. Connect a short (from your calibration kit) to one end of the cable to short the center conductor to the outer conductor. 2. While wearing a grounded wrist strap, grasp the outer shell of the cable connector. 3. Connect the other end of the cable to the test port. 4. Remove the short from the cable.

Refer to Chapter 6 , “Replaceable Parts,” for part numbers and instructions for ordering ESD protective devices.

Figure 3-1 ESD Protection Setup

Agilent Technologies E — Figure 3-1 ESD Protection Setup diagram (page 18)
Fig. 3 · Figure 3-1 ESD Protection Setup
Page 19 of the Agilent Technologies E user manual — Inspect the Mating Plane Surfaces
Page 19 · click to enlarge
Page 19

Use, Maintenance, and Care of the Devices

Visual Inspection

Visual inspection and, if necessary, cleaning should be done every time a connection is made. Metal particles from the connector threads may fall into the connector when it is disconnected. One connection made with a dirty or damaged connector can damage both connectors beyond repair.

In some cases, magnification is necessary to see damage on a connector; a magnifying device with a magnification of ≥10× is recommended. However, not all defects that are visible only under magnification will affect the electrical performance of the connector. Use the following guidelines when evaluating the integrity of a connector.

Look for Obvious Defects and Damage First

Examine the connectors first for obvious defects and damage: badly worn plating on the connector interface, deformed threads, or bent, broken, or misaligned center conductors. Connector nuts should move smoothly and be free of burrs, loose metal particles, and rough spots.

What Causes Connector Wear?

Connector wear is caused by connecting and disconnecting the devices. The more use a connector gets, the faster it wears and degrades. The wear is greatly accelerated when connectors are not kept clean, or are connected incorrectly.

Connector wear eventually degrades performance of the device. Calibration devices should have a long life if their use is on the order of a few times per week. Replace devices with worn connectors.

The test port connectors on the network analyzer test set may have many connections each day, and are therefore more subject to wear. It is recommended that an adapter be used as a test port saver to minimize the wear on the test set’s test port connectors.

Inspect the Mating Plane Surfaces

Flat contact between the connectors at all points on their mating plane surfaces is required for a good connection. Look especially for deep scratches or dents, and for dirt and metal particles on the connector mating plane surfaces. Also look for signs of damage due to excessive or uneven wear or misalignment.

Light burnishing of the mating plane surfaces is normal, and is evident as light scratches or shallow circular marks distributed more or less uniformly over the mating plane surface. Other small defects and cosmetic imperfections are also normal. None of these affect electrical or mechanical performance.

If a connector shows deep scratches or dents, particles clinging to the mating plane surfaces, or uneven wear, clean and inspect it again. Devices with damaged connectors should be discarded. Determine the cause of damage before connecting a new, undamaged connector in the same configuration.

Visual Inspection

Page 20 of the Agilent Technologies E user manual — Cleaning Connectors
Page 20 · click to enlarge
Page 20

Use, Maintenance, and Care of the Devices Cleaning Connectors

Inspect the Precision Slotless Connectors (female)

Precision slotless female connectors are used to improve accuracy. The slotless contacts are not affected by the slight variations in male contact pin diameter. However, it is still advisable to inspect them regularly for damage.

Cleaning Connectors

Clean connectors are essential for ensuring the integrity of RF and microwave coaxial connections.

1. Use Compressed Air or Nitrogen

Use compressed air (or nitrogen) to loosen particles on the connector mating plane surfaces.

You can use any source of clean, dry, low-pressure compressed air or nitrogen that has an effective oil-vapor filter and liquid condensation trap placed just before the outlet hose.

Ground the hose nozzle to prevent electrostatic discharge, and set the air pressure to less than 414 kPa (60 psi) to control the velocity of the air stream. High-velocity streams of compressed air can cause electrostatic effects when directed into a connector. These electrostatic effects can damage the device. Refer to “Electrostatic Discharge” earlier in this chapter for additional information.

2. Clean the Connector Threads

Use isopropyl alcohol with adequate ventilation and avoid contact with eyes, skin, and clothing. It causes skin irritation, may cause eye damage, and is harmful if swallowed or inhaled. It may be harmful if absorbed through the skin. Wash thoroughly after handling.

In case of spill, soak up with sand or earth. Flush spill area with water.

Dispose of isopropyl alcohol in accordance with all applicable federal, state, and local environmental regulations.

Page 21 of the Agilent Technologies E user manual — Inspect the Precision Slotless Connectors (female)
Page 21 · click to enlarge
Page 21

Use, Maintenance, and Care of the Devices

Use a lint-free swab or cleaning cloth moistened with isopropyl alcohol to remove any dirt or stubborn contaminants on a connector that cannot be removed with compressed air or nitrogen. Refer to Table 6-3 on page 6-5 for part numbers for isopropyl alcohol and cleaning swabs.

a. Apply a small amount of isopropyl alcohol to a lint-free cleaning swab.

b. Clean the connector threads.

c. Let the alcohol evaporate, then blow the threads dry with a gentle stream of clean,

low-pressure compressed air or nitrogen. Always completely dry a connector before you reassemble or use it.

3. Clean the Mating Plane Surfaces

a. Apply a small amount of isopropyl alcohol to a lint-free cleaning swab.

b. Clean the center and outer conductor mating plane surfaces. When cleaning a female

connector, avoid snagging the swab on the center conductor contact fingers by using short strokes.

c. Let the alcohol evaporate, then blow the connector dry with a gentle stream of clean,

low-pressure compressed air or nitrogen. Always completely dry a connector before you reassemble or use it.

4. Reinspect

Inspect the connector again to make sure that no particles or residue are present.

Cleaning Connectors

Page 22 of the Agilent Technologies E user manual — Final Connection Using a Torque Wrench
Page 22 · click to enlarge
Page 22

Use, Maintenance, and Care of the Devices Connections

Connections

Good connections require a skilled operator. The most common cause of measurement error is bad connections. The following procedures illustrate how to make good connections.

How to Make a Connection

Preliminary Connection

1. Ground yourself and all devices. Wear a grounded wrist strap and work on a grounded, conductive table mat. Refer to “Electrostatic Discharge” on page 3-2 for ESD precautions.
2. Visually inspect the connectors. Refer to “Visual Inspection” on page 3-3 .
3. If necessary, clean the connectors. Refer to “Cleaning Connectors” on page 3-4 .
4. Carefully align the connectors. The male connector center pin must slip concentrically into the contact finger of the female connector.
5. Push the connectors straight together.

Do not twist or screw the connectors together. As the center conductors mate, there is usually a slight resistance.

6. The preliminary connection is tight enough when the mating plane surfaces make uniform, light contact. Do not overtighten this connection.

A connection in which the outer conductors make gentle contact at all points on both mating surfaces is sufficient. Very light finger pressure is enough to accomplish this.

7. Make sure the connectors are properly supported. Relieve any side pressure on the connection from long or heavy devices or cables.

Final Connection Using a Torque Wrench

1. Use a torque wrench to make a final connection. Table 3-1 provides information about the torque wrench recommended for use with this calibration kit. A torque wrench is not included in the calibration kit. Refer to Chapter 6 for part number and ordering information.
Page 23 of the Agilent Technologies E user manual — Figure 3-2 Wrench Positions
Page 23 · click to enlarge
Page 23

Use, Maintenance, and Care of the Devices

Table 3-1 Torque Wrench Information

Connector Type Torque Setting Torque Tolerance

Type-N 135 N-cm (12 in-lb) ±13.5 N-cm (±1.2 in-lb)

Using a torque wrench guarantees that the connection is not too tight, preventing possible connector damage. It also guarantees that all connections are equally tight each time.

2. Prevent the rotation of anything other than the connector nut that you are tightening. It may be possible to do this by hand if one of the connectors is fixed (as on a test port). In all situations, however, it is recommended that you use an open-end wrench to keep the body of the device from turning. Refer to Chapter 6 “Replaceable Parts” for part number and ordering information.
3. Position both wrenches within 90 degrees of each other before applying force. See Figure 3-2 . Wrenches opposing each other (greater than 90 degrees apart) will cause a lifting action which can misalign and stress the connections of the devices involved. This is especially true when several devices are connected together.

Figure 3-2 Wrench Positions

Connections

Page 24 of the Agilent Technologies E user manual — Figure 3-3 Using the Torque Wrench
Page 24 · click to enlarge
Page 24

Use, Maintenance, and Care of the Devices Connections

4. Hold the torque wrench lightly, at the end of the handle only (beyond the groove). See Figure 3-3 .

Figure 3-3 Using the Torque Wrench

5. Apply downward force perpendicular to the wrench handle. See Figure 3-3 . This applies torque to the connection through the wrench.

Do not hold the wrench so tightly that you push the handle straight down along its length rather than pivoting it, otherwise you apply an unknown amount of torque.

6. Tighten the connection just to the torque wrench break point. The wrench handle gives way at its internal pivot point. See Figure 3-3 . Do not tighten the connection further.

Do not pivot the wrench handle on your thumb or other fingers, otherwise you apply an unknown amount of torque to the connection when the wrench reaches its break point.

Do not twist the head of the wrench relative to the outer conductor mating plane. If you do, you apply more than the recommended torque.

Page 25 of the Agilent Technologies E user manual — How to Separate a Connection
Page 25 · click to enlarge
Page 25

Use, Maintenance, and Care of the Devices

Connecting and Disconnecting the Two-Piece Female Open (85036B)

The female open standard in the 85036B calibration kit is composed of two parts: the open body (outer conductor) and the center conductor extender. Refer to Figure 3-4.

Figure 3-4 Connecting the Two-Piece Female Open

To connect the female open:

1. Connect the open body to the male test port.
2. Insert the center conductor extender into the hole at the end of the body and push gently until the center conductors mate.

To disconnect the female open:

1. Remove the center conductor extender by pulling gently outwards without twisting, rocking, or bending the extender or the body.
2. Disconnect the body from the test port.

How to Separate a Connection

To avoid lateral (bending) force on the connector mating plane surfaces, always support the devices and connections.

1. Use an open-end wrench to prevent the device body from turning.
2. Use another open-end wrench to loosen the connector nut.
3. Complete the separation by hand, turning only the connector nut.
4. Pull the connectors straight apart without twisting, rocking, or bending either of the connectors.

Connections

Page 26 of the Agilent Technologies E user manual — Handling and Storage
Page 26 · click to enlarge
Page 26

Use, Maintenance, and Care of the Devices Handling and Storage

Handling and Storage

  • Install the protective end caps and store the calibration devices in the foam-lined storage case when not in use.
  • Never store connectors loose in a box, desk, or bench drawer. This is the most common cause of connector damage during storage.
  • Keep connectors clean.
  • Do not touch mating plane surfaces. Natural skin oils and microscopic particles of dirt are easily transferred to a connector interface and are very difficult to remove.
  • Do not set connectors contact-end down on a hard surface. The plating and the mating plane surfaces can be damaged if the interface comes in contact with any hard surface.
Page 27 of the Agilent Technologies E user manual — 4 Performance Verification
Page 27 · click to enlarge
Page 27

4 Performance Verification

Page 28 of the Agilent Technologies E user manual — How Agilent Verifies the Devices in This Kit
Page 28 · click to enlarge
Page 28

Introduction

The performance of your calibration kit can only be verified by returning the kit to Agilent Technologies for recertification. The equipment required to verify the specifications of the devices in the kit has been specially manufactured and is not commercially available.

How Agilent Verifies the Devices in This Kit

Agilent verifies the specifications of these devices as follows:

1. The residual microwave error terms of the test system are verified with precision airlines and shorts that are directly traced to NIST (National Institute of Standards and Technology). The airline and short characteristics are developed from mechanical measurements. The mechanical measurements and material properties are carefully modeled to give very accurate electrical representation. The mechanical measurements are then traced to NIST through various plug and ring gages and other mechanical measurements.
2. Each calibration device is electrically tested on this system. For the initial (before sale) testing of the calibration devices, Agilent includes the test measurement uncertainty as a guardband to guarantee each device meets the published specification. For recertifications (after sale), no guardband is used and the measured data is compared directly with the specification to determine the pass or fail status. The measurement uncertainty for each device is, however, recorded in the calibration report that accompanies recertified kits.

These two steps establish a traceable link to NIST for Agilent to the extent allowed by the institute’s calibration facility. The specifications data provided for the devices in this kit is traceable to NIST through Agilent Technologies.

Performance Verification Introduction

Page 29 of the Agilent Technologies E user manual — Where to Send a Kit for Recertification
Page 29 · click to enlarge
Page 29

Recertification

The following will be provided with a recertified kit:

  • a new calibration sticker affixed to the case
  • a certificate of calibration
  • a calibration report for each device in the kit listing measured values, specifications, and uncertainties

Agilent Technologies offers a Standard calibration for the recertification of this kit. For more information, contact Agilent Technologies. See Table 5-1 on page 5-3.

Limited Recertification

The 75Ω loads in the calibration kit are specified for use up to 3 GHz. For many applications, the performance above 2 GHz is not utilized.

For a standard recertification, the devices in the calibration kit are tested and calibrated up to 3 GHz. However, a limited recertification can be requested. For this limited recertification, the devices are tested and calibrated up to 2 GHz.

To request a limited recertification, make sure the following is clearly written on the order: Limited Calibration DC–2 GHz.

All loads that receive a limited calibration are supplied with a limited calibration label applied to the device.

How Often to Recertify

The suggested initial interval for recertification is 12 months or sooner. The actual need for recertification depends on the use of the kit. After reviewing the results of the initial recertification, you may establish a different recertification interval that reflects the usage and wear of the kit.

Where to Send a Kit for Recertification

Contact Agilent Technologies for information on where to send your kit for recertification. See Table 5-1 on page 5-3.

When you return the kit, complete and attach a service tag. Refer to “Returning a Kit or Device to Agilent” on page 5-3 for details.

Performance Verification · Recertification

Page 30 of the Agilent Technologies E user manual — Recertification
Page 30 · click to enlarge
Page 30

Performance Verification Recertification

Page 31 of the Agilent Technologies E user manual — 5 Troubleshooting
Page 31 · click to enlarge
Page 31

5 Troubleshooting

Page 32 of the Agilent Technologies E user manual — Figure 5-1 Troubleshooting Flowchart
Page 32 · click to enlarge
Page 32

Troubleshooting Process

If you suspect a bad calibration, or if your network analyzer does not pass performance verification, follow the steps in Figure 5-1.

Figure 5-1 Troubleshooting Flowchart

Agilent Technologies E — Figure 5-1 Troubleshooting Flowchart diagram (page 32)
Fig. 4 · Figure 5-1 Troubleshooting Flowchart

Troubleshooting Troubleshooting Process

Page 33 of the Agilent Technologies E user manual — Table 5-1 Contacting Agilent
Page 33 · click to enlarge
Page 33

Troubleshooting Returning a Kit or Device to Agilent

Returning a Kit or Device to Agilent

If your kit or device requires service, contact the Agilent Technologies office nearest you for information on where to send it. See Table 5-1. Include a service tag (located near the end of this manual) on which you provide the following information:

  • your company name and address
  • a technical contact person within your company, and the person's complete phone number
  • the model number and serial number of the kit
  • the part number and serial number of each device
  • the type of service required
  • a detailed description of the problem and how the device was being used when the problem occurred (such as calibration or measurement)

Table 5-1 Contacting Agilent

Online assistancewww.agilent.com/find/assist

United States (tel) 1 800 452 4844

Latin America (tel) (305) 269 7500 (fax) (305) 269 7599

Canada (tel) 1 877 894 4414 (fax) (905) 282-6495

Europe (tel) (+31) 20 547 2323 (fax) (+31) 20 547 2390

New Zealand (tel) 0 800 738 378 (fax) (+64) 4 495 8950

Japan (tel) (+81) 426 56 7832 (fax) (+81) 426 56 7840

Australia (tel) 1 800 629 485 (fax) (+61) 3 9210 5947

Singapore (tel) 1 800 375 8100 (fax) (65) 836 0252

Malaysia (tel) 1 800 828 848 (fax) 1 800 801 664

Philippines (tel) (632) 8426802 (tel) (PLDT subscriber only): 1 800 16510170 (fax) (632) 8426809 (fax) (PLDT subscriber only): 1 800 16510288

Thailand (tel) outside Bangkok(088) 226 008 (tel) within Bangkok: (662) 661 3999 (fax) (66) 1 661 3714

Hong Kong (tel) 800 930 871 (fax) (852) 2506 9233

Taiwan (tel) 0800-047-866 (fax) (886) 2 25456723

People’s Republic of China (tel) (preferred): 800-810-0189 (tel) (alternate): 10800-650-0021 (fax) 10800-650-0121

India (tel) 1-600-11-2929 (fax) 000-800-650-1101

Page 34 of the Agilent Technologies E user manual — Returning a Kit or Device to Agilent
Page 34 · click to enlarge
Page 34

Troubleshooting Returning a Kit or Device to Agilent

Page 35 of the Agilent Technologies E user manual — 6 Replaceable Parts
Page 35 · click to enlarge
Page 35

6 Replaceable Parts

Page 36 of the Agilent Technologies E user manual — Table 6-1 Replaceable Parts for the 85036B Calibration Kit
Page 36 · click to enlarge
Page 36

Introduction

Table 6-1 lists the replacement part numbers for items included in the 85036B calibration kit and Figure 6-1 illustrates each of these items.

Table 6-2 lists the replacement part numbers for items included in the 85036E calibration kit and Figure 6-2 illustrates each of these items.

Table 6-3 lists the replacement part numbers for items recommended or required for successful operation but not included in the calibration kit.

To order a listed part, note the description, the part number, and the quantity desired. Telephone or send your order to Agilent Technologies. See Table 5-1 on page 5-3.

Table 6-1 Replaceable Parts for the 85036B Calibration Kit

Description Qty Per Kit Agilent Part Number

Calibration Devices (75Ω Type-N)

Male broadband load 1 00909-60019

Female broadband load 1 00909-60020

Male short 1 85036-60012

Female short 1 85036-60011

Male open 1 85032-60007

Female open body 1 85032-20001

Female open center conductor extender 1 85036-60019

Type-N-male to male 1 85036-60013

Type-N-female to female 1 85036-60014

Type-N-male to female 1 85036-60015

Calibration Kit Storage Case

Foam pad set 1 85036-80002

User’s and service guide 1 85036-90016

Male protective end caps as required 1401-0214

Female protective end caps as required 1401-0225

Tube package (for center conductor extender) 1 1540-1027

Show 4 more labels / short text on this page
Replaceable Parts Introduction · Adapters · Box 1 85036-80001 · Miscellaneous Items
Page 37 of the Agilent Technologies E user manual — Figure 6-1 Replaceable Parts for the 85036B Calibration Kit
Page 37 · click to enlarge
Page 37

Figure 6-1 Replaceable Parts for the 85036B Calibration Kit

Agilent Technologies E — Figure 6-1 Replaceable Parts for the 85036B Calibration Kit diagram (page 37)
Fig. 5 · Figure 6-1 Replaceable Parts for the 85036B Calibrat…

Replaceable Parts · Introduction

Page 38 of the Agilent Technologies E user manual — Figure 6-2 Replaceable Parts for the 85036E Calibration Kit
Page 38 · click to enlarge
Page 38

Table 6-2 Replaceable Parts for the 85036E Calibration Kit

Description Qty Per Kit Agilent Part Number

Calibration Devices (75Ω Type-N)

Male broadband load 1 00909-60019

Male combination open/short 1 85036-60016

Calibration Kit Storage Case

Foam pad set 1 85023-80005

User’s and service guide 1 85036-90016

Male protective end cap as required

Figure 6-2 Replaceable Parts for the 85036E Calibration Kit

Agilent Technologies E — Figure 6-2 Replaceable Parts for the 85036E Calibration Kit diagram (page 38)
Fig. 6 · Figure 6-2 Replaceable Parts for the 85036E Calibrat…
Show 4 more labels / short text on this page
Replaceable Parts Introduction · Box 1 9211-1582 · Miscellaneous Items · 1401-0214
Page 39 of the Agilent Technologies E user manual — Introduction
Page 39 · click to enlarge
Page 39

Table 6-3 Replaceable Parts—Items Not Included in the 85036B and 85036E Calibration Kits

Description Qty Agilent Part Number

Connector Gages (75Ω Type-N)

Connector gage set 1 85036B K01

3/4 in, 135 N-cm (12 in-lb) torque wrench 1 8710-1766

1/2 in and 9/16 in open-end wrench 1 8710-1770

3/4 in open-end wrench 1 8720-0011

Grounding wrist strap 1 9300-1367

5 ft grounding cord for wrist strap 1 9300-0980

2 ft by 4 ft conductive table mat with 15 ft grounding wire 1 9300-0797

ESD heel strap 1 9300-1308

Isopropyl alcohol 30 ml 8500-5344

Foam tipped cleaning swabs 100 9301-1243

Connector care - quick reference card 1 08510-90360

Show 6 more labels / short text on this page
Replaceable Parts · Introduction · Wrenches · ESD Protective Devices · Connector Cleaning Supplies · Miscellaneous Items
Page 40 of the Agilent Technologies E user manual — Introduction
Page 40 · click to enlarge
Page 40

Replaceable Parts Introduction

Page 41 of the Agilent Technologies E user manual — A-1
Page 41 · click to enlarge
Page 41

A Standard Definitions

A-1

Page 42 of the Agilent Technologies E user manual — A-2 85036B/E
Page 42 · click to enlarge
Page 42

Standard Definitions Standard Class Assignments

Standard Class Assignments

Class assignment organizes calibration standards into a format compatible with the error models used in the measurement calibration. A class or group of classes corresponds to the systematic errors to be removed from the measured network analyzer response. Table A-1 lists the classes of the devices in this calibration kit.

Table A-1 Standard Class Assignments

Calibration Kit LabelN 75 Ω

Class A B C D E F G Standard Class Label

S11A 2 8 Opens

S11B 1 7 Shorts

S11C 3 Load

S22A 2 8 Opens

S22B 1 7 Shorts

S22C 3 Load

Forward Transmission 4 Thru

Reverse Transmission 4 Thru

Forward Match 4 Thru

Reverse Match 4 Thru

Response 1 7 2 8 4 Response

Response and Isolation 1 7 2 8 4 Response & Isolation

A-2 85036B/E

Page 43 of the Agilent Technologies E user manual — S 11 A
Page 43 · click to enlarge
Page 43

Standard Definitions Standard Class Assignments

Blank Form

The standard class assignments listed in Table A-1 may be changed to meet your specific requirements. Table A-2 is provided to record the modified standard class assignments.

Table A-2 Standard Class Assignments Blank Form

Calibration Kit Label______________

Class A B C D E F G Standard Class Label

S 11 A

S 11 B

S 11 C

S 22 A

S 22 B

S 22 C

85036B/E A-3

Show 6 more labels / short text on this page
Forward Transmission · Reverse Transmission · Forward Match · Reverse Match · Response · Response and Isolation
Page 44 of the Agilent Technologies E user manual — A-4 85036B/E
Page 44 · click to enlarge
Page 44

Standard Definitions Nominal Standard Definitions

Nominal Standard Definitions

Standard definitions provide the constants needed to mathematically model the electrical characteristics (delay, attenuation, and impedance) of each calibration standard. The nominal values of these constants are theoretically derived from the physical dimensions and material of each calibration standard, or from actual measured response. These values are used to determine the measurement uncertainties of the network analyzer. The standard definitions in Table A-3 list typical calibration kit parameters used to specify the mathematical model of each device. This information must be loaded into the network analyzer to perform valid calibrations. Refer to your network analyzer’s user’s guide for instructions on loading calibration constants.

Setting the System Impedance

This kit contains only 75 ohm devices. Ensure the system impedance (Z0) is set to 75 ohms. Refer to your network analyzer’s user’s guide for instructions on setting system impedance.

A-4 85036B/E

Page 45 of the Agilent Technologies E user manual — 85036B/E A-5
Page 45 · click to enlarge
Page 45

Standard Definitions Nominal Standard Definitions

Table A-3 Standard Definitions

a = 75 Ω Calibration Kit Label: N 75 Ω

C3 ×10−45 F/Hz3

C1 ×10−27 F/Hz

C2 ×10−36 F/Hz2

Fixed or Slidingc

C0 × 10 − 15 F

Loss (Ω/s)

Z 0 Ω

1 Short 0 75 1.13G 0 999 Coax Short (m)d

2 Open 63.5 84 56 0 0 75 1.13G 0 999 Coax Open (m)d

3 Load Fixed 0 75 1.13G 0 999 Coax Broad- band

0 75 1.13G 0 999 Coax Thru

7 Short 17.544 75 1.13G 0 999 Coax Short (f)d

8 Open 41e 40 5 0 17.544 75 1.13G 0 999 Coax Open (f)d

a. Ensure system impedance (Z0) of network analyzer is set to this value. b. Open, short, load, delay/thru, or arbitrary impedance. c. Load or arbitrary impedance only. d. Standard labels that specify sex, (m) or (f), refer to the sex of the analyzer’s test port connector. e. If your instrument is an Agilent 8752B or 8711 Option 1EC, and you are calibrating at the test port, this number is 33. For all other instruments and applications, including calibrating at a point other than the test port (such as the end of a cable), 41 is the correct number. This is because the 8752B and 8711 Option 1EC have special female test ports with longer, more rugged fingers.

85036B/E A-5

Show 11 more labels / short text on this page
System Z0 · Coax or Waveguide · Standardb · Offset Freq (GHz) · Standard Labeld · Delay (ps) · Number · Type · Min · Max · 4 Delay/ Thru
Page 46 of the Agilent Technologies E user manual — A-6 85036B/E
Page 46 · click to enlarge
Page 46

Standard Definitions Nominal Standard Definitions

Blank Form

The standard definitions listed in Table A-3 may be changed to meet your specific requirements. Table A-4 is provided to record the modified standard definitions.

Table A-4 Standard Definitions Blank Form

Calibration Kit Label____________________

C3 ×10−45 F/Hz3

C1 ×10−27 F/Hz

C2 ×10−36 F/Hz2

Fixed or Slidingc

C0 × 10 − 15 F

Loss (GΩ/s)

Z 0 Ω

a. Ensure system Z0 of network analyzer is set to this value.

b. Open, short, load, delay/thru, or arbitrary impedance.

c. Load or arbitrary impedance only.

d. Standard labels that specify sex, (m) or (f), refer to the sex of the analyzer’s test port connector.

A-6 85036B/E

Show 11 more labels / short text on this page
a = ______________ · System Z0 · Standardb · Offset Freq (GHz) · Coax or Waveguide · Standard Label d · Delay (ps) · Number · Type · Min · Max
Page 48 of the Agilent Technologies E user manual — Index
Page 48 · click to enlarge
Page 48

Index

when making connections, 3-6 environmental requirements, 2-2 equipment

limited calibration, 4-3 limited calibration

requesting, 4-3 loads

but not supplied, 1-3, 6-2, 6-5 supplied, 1-2, 6-2, 6-4 ESD, 3-2

M maintenance, 3-2

precautions, 3-2, 3-4 supplies

preventive, 1-8 maintenance of devices, 1-8 making connections, 3-6 manual

part numbers, 6-5 when making connections, 3-6

F female open, 3-9

part number, 6-5 mating plane

connecting, 3-9 disconnecting, 3-9 frequency

connector, 3-5 mating plane inspection, 3-3 mating plane surfaces, 3-5 mechanical specifications, 2-3

specifications, 2-4 frequency range, 2-4, 4-3

N nitrogen, 3-4 nominal standard definitions, A-4 numbers

master, 6-5 gage master

part numbers, 6-5 gages, 1-3

sales and service, 5-3 open-end wrench, 1-3, 3-9

H handling, 3-10 history, calibration kit, 1-7 humidity, 2-2

part number, 6-5 opens, 1-2 opens and shorts, 1-2 ordering parts, 6-2

P part numbers, 6-2 parts

system, A-4 incoming inspection, 1-4 inside diameter of outer conductor, specifications, 2-3 inspection

ordering, 6-2 replaceable, 6-2 required but not supplied, 6-2 performance verification, 4-2

damage, 3-3 defects, 3-3 incoming, 1-4 mating plane, 3-3 visual, 3-3 isopropyl alcohol

fail, 5-2 permanently stored calibration definitions, 1-3 pin depth

typical values, 2-3 precision slotless connectors, 3-4 preventive maintenance, 1-8

as cleaning solvent, 3-4

R recertification, 4-3

contents, 1-2 overview, 1-2 return to Agilent, 5-3 serial number, 1-6

how often?, 4-3 interval, 4-3 limited, 4-3 what’s included, 4-3 where to send your kit, 4-3 regulations

environmental, 3-5 replaceable parts, 6-2

part number, 6-5

Show 12 more labels / short text on this page
required, 1-3 · broadband, 1-2 · G gage · connector · serial, 1-6 · recording, 1-6 · O offices · I impedance · K kit · L label · calibration, 4-3 · 85036B/E Index-2
Page 49 of the Agilent Technologies E user manual — Index
Page 49 · click to enlarge
Page 49

Index

device, 2-2 error-corrected temperature range, 2-2 measurement, 2-2 operating range, 2-2 verification and measurement, 2-2 test data, 4-2 threads

environmental, 2-2 return loss

specifications, 2-4 return, kit or device, 5-3

connector, 3-4 torque wrench, 1-3, 3-6

and service offices, 5-3 sales and service offices, 5-3 serial numbers, 1-6

part number, 6-5 specifications, 3-6 traceability, 4-2 troubleshooting, 5-2 two-piece female open, 3-9

devices, 1-6 recording, 1-6 service, 5-2

and sales offices, 5-3 service tag, 1-4, 4-3, 5-3 sex

connector, 1-8 shipment

performance, 4-2 temperature, 2-2 visual inspection, 3-3

damage, 1-4 verifying complete, 1-4 shorts, 1-2 specifications, 2-2

W warranty, documentation, ii wrench

altitude, 2-2 certification, 2-4 distance from ref plane to shorting plane, 2-3 electrical, 2-4 environmental, 2-2 frequency, 2-4 humidity, 2-2 inside diameter of outer conductor, 2-3 mechanical, 2-3 return loss, 2-4 temperature, 2-2 verifying, 4-2 standard class assignments, A-2

open-end, 1-3, 3-9

part number, 6-5 undoing connections, 3-9 torque, 1-3

part number, 6-5 wrenches

part numbers, 6-5

blank form, A-3 standard definitions, A-4, A-6

blank form, A-6 nominal, A-4 standards

calibration, 2-4 NIST, 2-4, 4-3 static discharge, 3-2 sticker

calibration, 4-3 storage, 3-10

altitude, 2-2 humidity, 2-2 temperature, 2-2 supplies

cleaning, 1-3 system impedance

service, 1-4, 4-3, 5-3 temperature

Show 7 more labels / short text on this page
requirements · S sales · V verification · setting, A-4 · T tag · calibration, 2-2 · 85036B/E Index-3

Ask about the Agilent Technologies E

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the Agilent Technologies E

Be the first to ask — other owners or our team will answer.

Original document

Agilent Technologies E manual PDF

49 pages·809 KB·Work Light·PDF
Click any page above to open it here
Page 1 / …
100%
Loading pages…