Agilent Technologies 8753D

678 pages·4.8 MB·Network Router·PDF·ENDEFRITNLJA
Agilent Technologies 8753D
Table of Contents 21 sections NoticeHP to Agilent Technologies Transition · Title Page8753D Network Analyzer User's Guide · Contacting Agilent Technologies · User's Guide Overview · … ShowHide
Page 1 of the Agilent Technologies 8753D user manual — Printed in USA March 2000
Page 1 · click to enlarge
Page 1

Notice

Hewlett-Packard to Agilent Technologies Transition

This documentation supports a product that previously shipped under the Hewlett- Packard company brand name. The brand name has now been changed to Agilent Technologies. The two products are functionally identical, only our name has changed. The document still includes references to Hewlett-Packard products, some of which have been transitioned to Agilent Technologies.

Agilent Technologies 8753D — Notice diagram (page 1)
Fig. 1 · Notice

Printed in USA March 2000

Page 2 of the Agilent Technologies 8753D user manual — ABCDE
Page 2 · click to enlarge
Page 2

HP 8753D Network Analyzer

ABCDE

HP P art No . 08753-90257 Supersedes October 1997

Printed in US A December 1997

User's Guide

Page 3 of the Agilent Technologies 8753D user manual — Title Page8753D Network Analyzer User's Guide
Page 3 · click to enlarge
Page 3

The information contained in this document is subject to change without notice .

Hewlett-P ackard makes no warranty of any kind with regard to this material, including

but not limited to , the implied warranties of merchantability and tness for a particular

purpose . Hewlett-P ackard shall not be liable for errors contained herein or for incidental or

consequential damages in connection with the furnishing, performance , or use of this material.

c Copyright Hewlett-P ackard Company 1994, 1995, 1997

All Rights Reserved. Reproduction, adaptation, or translation without prior written permission

is prohibited, except as allowed under the copyright laws .

1400 F ountaingrove P arkway , Santa Rosa, CA 95403-1799, USA

Page 4 of the Agilent Technologies 8753D user manual — Title Page8753D Network Analyzer User's Guide
Page 4 · click to enlarge
Page 4

Hewlett-P ackard Company certi es that this product met its published speci cations at the

time of shipment from the factory . Hewlett-P ackard further certi es that its calibration

measurements are traceable to the United States National Institute of Standards and

T echnology , to the extent allowed by the Institute's calibration facility , and to the calibration

facilities of other International Standards Organization members .

This Hewlett-P ackard instrument product is warranted against defects in material and

workmanship for a period of one year from date of shipment. During the warranty period,

Hewlett-P ackard Company will, at its option, either repair or replace products which prove to

be defective .

F or warranty service or repair , this product must be returned to a service facility designated by

Hewlett-P ackard. Buyer shall prepay shipping charges to Hewlett-P ackard and Hewlett-P ackard

shall pay shipping charges to return the product to Buyer . However , Buyer shall pay all

shipping charges , duties , and taxes for products returned to Hewlett-P ackard from another

country .

Hewlett-P ackard warrants that its software and rmware designated by Hewlett-P ackard for

use with an instrument will execute its programming instructions when properly installed on

that instrument. Hewlett-P ackard does not warrant that the operation of the instrument, or

software , or rmware will be uninterrupted or error-free .

Limit a tion of W arranty

The foregoing warranty shall not apply to defects resulting from improper or inadequate

maintenance by Buyer , Buyer-supplied software or interfacing, unauthorized modi cation or

misuse , operation outside of the environmental speci cations for the product, or improper

site preparation or maintenance .

NO OTHER W ARRANTY IS EXPRESSED OR IMPLIED . HEWLETT-P A CKARD SPECIFICALL Y

DISCLAIMS THE IMPLIED W ARRANTIES OF MERCHANT ABILITY AND FITNESS FOR A

P ARTICULAR PURPOSE.

THE REMEDIES PRO VIDED HEREIN ARE BUYER'S SOLE AND EXCLUSIVE REMEDIES.

HEWLETT-P A CKARD SHALL NOT BE LIABLE FOR ANY DIRECT , INDIRECT , SPECIAL,

INCIDENT AL, OR CONSEQUENTIAL D AMA GES, WHETHER B ASED ON CONTRA CT , TORT ,

OR ANY OTHER LEGAL THEORY .

Certi cation · W arranty · Ex clusive Remedies

Page 5 of the Agilent Technologies 8753D user manual — Title Page8753D Network Analyzer User's Guide
Page 5 · click to enlarge
Page 5

Clean the cabinet, using a damp cloth only .

Product maintenance agreements and other customer assistance agreements are available for

Hewlett-P ackard products .

F or any assistance , contact your nearest Hewlett-P ackard Sales and Service Oce .

Maintenance · Assistance

Page 6 of the Agilent Technologies 8753D user manual — Table 1-1 Contacting Agilent
Page 6 · click to enlarge
Page 6

Contacting Agilent

By internet, phone, or fax, get assistance with all your test and measurement needs.

Table 1-1 Contacting Agilent

Online assistance: www.agilent.com/find/assist

United States (tel) 1 800 452 4844

Latin America (tel) (305) 269 7500 (fax) (305) 269 7599

Canada (tel) 1 877 894 4414 (fax) (905) 282-6495

Europe (tel) (+31) 20 547 2323 (fax) (+31) 20 547 2390

New Zealand (tel) 0 800 738 378 (fax) (+64) 4 495 8950

Japan (tel) (+81) 426 56 7832 (fax) (+81) 426 56 7840

Australia (tel) 1 800 629 485 (fax) (+61) 3 9210 5947

Asia Call Center Numbers

Country Phone Number Fax Number

Singapore 1-800-375-8100 (65) 836-0252

Malaysia 1-800-828-848 1-800-801664

Philippines (632) 8426802 1-800-16510170 (PLDT Subscriber Only)

(632) 8426809 1-800-16510288 (PLDT Subscriber Only)

Thailand (088) 226-008 (outside Bangkok) (662) 661-3999 (within Bangkok)

(66) 1-661-3714

Hong Kong 800-930-871 (852) 2506 9233

Taiwan 0800-047-866 (886) 2 25456723

People’s Republic of China

800-810-0189 (preferred) 10800-650-0021 10800-650-0121

India 1-600-11-2929 000-800-650-1101

2 Chapter1

Page 7 of the Agilent Technologies 8753D user manual — Contacting Agilent Technologies
Page 7 · click to enlarge
Page 7

The following safety symbols are used throughout this manual. F amiliarize yourself with each

of the symbols and its meaning before operating this instrument.

correctly performed or adhered to , would result in damage to or destruction

of the instrument. Do not proceed beyond a caution note until the indicated

conditions are fully understood and met.

W arning denotes a hazard. It calls attention to a procedure which, if not

correctly performed or adhered to , could result in injury or loss of life .

Do not proceed beyond a warning note until the indicated conditions are

fully understood and met.

L The instruction documentation symbol. The product is marked with this symbol when it

is necessary for the user to refer to the instructions in the documentation.

\CE" The CE mark is a registered trademark of the European Community . (If accompanied by

a year , it is when the design was proven.)

\ISM1-A" This is a symbol of an Industrial Scienti c and Medical Group 1 Class A product.

\CSA" The CSA mark is a registered trademark of the Canadian Standards Association.

Safety Symbols · W arning · Instrument Markings

Page 8 of the Agilent Technologies 8753D user manual — Contacting Agilent Technologies
Page 8 · click to enlarge
Page 8

This is a Safety Class I product (provided with a protective earthing

ground incorporated in the power cord). The mains plug shall only be

inserted in a socket outlet provided with a protective earth contact. Any

interruption of the protective conductor , inside or outside the instrument,

is likely to make the instrument dangerous . Intentional interruption is

prohibited.

No operator serviceable parts inside . Refer servicing to quali ed

personnel. T o prevent electrical shock, do not remove covers .

Before switching on this instrument, make sure that the line voltage selector

switch is set to the voltage of the power supply and the correct fuse is

installed.

The opening of covers or removal of parts is likely to expose dangerous

voltages . Disconnect the instrument from all voltage sources while it is

being opened.

The power cord is connected to internal capacitors that may remain live

for 10 seconds after disconnecting the plug from its power supply .

F or continued protection against re hazard replace line fuse only with

same type and rating (F 3A/250V). The use of other fuses or material is

prohibited.

If this instrument is used in a manner not speci ed by Hewlett-P ackard

Co ., the protection provided by the instrument may be impaired.

This instrument has been designed and tested in accordance with IEC

Publication 348, Safety Requirements for Electronics Measuring Apparatus , and

has been supplied in a safe condition. This instruction documentation contains

information and warnings which must be followed by the user to ensure safe

operation and to maintain the instrument in a safe condition.

Show 7 more labels / short text on this page
General Safety Considerations · W arning · W arning · W arning · W arning · W arning · W arning
Page 9 of the Agilent Technologies 8753D user manual — User's Guide Overview
Page 9 · click to enlarge
Page 9

Chapter 1, \HP 8753D Description and Options ," describes features , functions , and available

options .

Chapter 2, \Making Measurements ," contains step-by-step procedures for making

measurements or using particular functions .

Chapter 3, \Making Mixer Measurements ," contains step-by-step procedures for making

calibrated and error-corrected mixer measurements .

Chapter 4, \Printing, Plotting, and Saving Measurement Results ," contains instructions

for saving to disk or the analyzer internal memory , and printing and plotting displayed

measurements .

Chapter 5, \Optimizing Measurement Results ," describes techniques and functions for

achieving the best measurement results .

Chapter 6, \Application and Operation Concepts ," contains explanatory-style information

about many applications and analyzer operation.

Chapter 7, \Speci cations and Measurement Uncertainties ," de nes the performance

capabilities of the analyzer .

Chapter 8, \Menu Maps ," shows softkey menu relationships .

Chapter 9, \Key De nitions ," describes all the front panel keys , softkeys , and their

corresponding HP-IB commands .

Chapter 10, \Error Messages ," provides information for interpreting error messages .

Chapter 11, \Compatible P eripherals ," lists measurement and system accessories , and

other applicable equipment compatible with the analyzer . Procedures for con guring the

peripherals , and an HP-IB programming overview are also included.

Chapter 12, \Preset State and Memory Allocation," contains a discussion of memory

allocation, memory storage , instrument state de nitions , and preset conditions .

Appendix A, \The CITI le Data F ormat and Key W ord Reference ," contains information on

the CITI le data format as well as a list of CITI le keywords .

User's Guide Overview

Page 10 of the Agilent Technologies 8753D user manual — User's Guide Overview
Page 10 · click to enlarge
Page 10

Network Analyzer Documentation Set

The Installation and Quick Start Guide

Agilent Technologies 8753D — User's Guide Overview diagram (page 10)
Fig. 2 · User's Guide Overview

familiarizes you with the network analyzer's

front and rear panels , electrical and

environmental operating requirements , as well

as procedures for installing, con guring, and

verifying the operation of the analyzer .

The User's Guide shows how to make

measurements , explains commonly-used

Agilent Technologies 8753D — User's Guide Overview diagram (page 10)
Fig. 3 · User's Guide Overview

features , and tells you how to get the most

performance from your analyzer .

The Quick Reference Guide provides a

summary of selected user features .

Agilent Technologies 8753D — User's Guide Overview diagram (page 10)
Fig. 4 · User's Guide Overview

The Programmer's Guide provides

programming information including an HP-IB

programming and command reference as well

as programming examples .

The System V eri cation and T est Guide

provides the system veri cation and

performance tests and the P erformance T est

Record for your analyzer .

Page 11 of the Agilent Technologies 8753D user manual — User's Guide Overview
Page 11 · click to enlarge
Page 11
Agilent Technologies 8753D — User's Guide Overview diagram (page 11)
Fig. 5 · User's Guide Overview
Page 13 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNN
Page 13 · click to enlarge
Page 13

T o Title the A ctive Channel Display 2-8

Using Analyzer Display Markers 2-9

T o Use Continuous and Discrete Markers 2-9

T o A ctivate Display Markers 2-10

T o Use Delta () Markers 2-11

T o A ctivate a Fixed Marker 2-11

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REF=FIXED MKR Key to activate a Fixed Reference Marker . . 2-12

NNNNNNNNNNNNNNNNNNNNNNNNNN

MKR ZERO Key to A ctivate a Fixed Reference Marker 2-13

T o Couple and Uncouple Display Markers 2-14

T o Use P olar F ormat Markers 2-14

T o Use Smith Chart Markers 2-15

T o Set Measurement P arameters Using Markers 2-16

Setting the Start Frequency 2-17

Setting the Stop Frequency 2-17

Setting the Center Frequency 2-18

Setting the Frequency Span 2-19

Setting the Display Reference V alue 2-20

Setting the Electrical Delay 2-21

Setting the CW Frequency 2-21

T o Search for a Speci c Amplitude 2-22

Searching for the Maximum Amplitude 2-22

Searching for the Minimum Amplitude 2-23

Searching for a T arget Amplitude 2-24

Searching for a Bandwidth 2-25

Tracking the Amplitude that Y ou are Searching 2-25

T o Calculate the Statistics of the Measurement Data 2-26

Measuring Magnitude and Insertion Phase Response 2-27

Measuring the Magnitude Response 2-27

Measuring Insertion Phase Response 2-28

Measuring Electrical Length and Phase Distortion 2-30

Measuring Electrical Length 2-30

Measuring Phase Distortion 2-32

Deviation From Linear Phase 2-32

Group Delay 2-33

T esting A Device with Limit Lines 2-36

Setting Up the Measurement P arameters 2-36

Creating Flat Limit Lines 2-37

Creating a Sloping Limit Line 2-39

Creating Single P oint Limits 2-41

Editing Limit Segments 2-42

Deleting Limit Segments 2-42

Running a Limit T est 2-43

Reviewing the Limit Line Segments 2-43

A ctivating the Limit T est 2-43

O setting Limit Lines 2-44

Measuring Gain Compression 2-45

Measuring Gain and Reverse Isolation Simultaneously 2-49

Measurements Using the Tuned Receiver Mode 2-51

Typical test setup 2-51

Tuned receiver mode in-depth description 2-51

Frequency Range 2-51

Compatible Sweep Types 2-51

External Source Requirements 2-52

T est Sequencing 2-53

Using the · Using the · Contents-2

Page 14 of the Agilent Technologies 8753D user manual — Table of Contents
Page 14 · click to enlarge
Page 14

Creating a Sequence 2-54

Running a Sequence 2-55

Stopping a Sequence 2-55

Editing a Sequence 2-56

Deleting Commands 2-56

Inserting a Command 2-56

Modifying a Command 2-57

Clearing a Sequence from Memory 2-57

Changing the Sequence Title 2-58

Naming Files Generated by a Sequence 2-58

Storing a Sequence on a Disk 2-59

Loading a Sequence from Disk 2-60

Purging a Sequence from Disk 2-60

Printing a Sequence 2-60

Cascading Multiple Example Sequences 2-61

Loop Counter Example Sequence 2-62

Generating Files in a Loop Counter Example Sequence 2-63

Limit T est Example Sequence 2-64

Measuring Swept Harmonics 2-66

Measuring a Device in the Time Domain (Option 010 Only) 2-68

Transmission Response in Time Domain 2-68

Re ection Response in Time Domain 2-73

Non-coaxial Measurements 2-76

3. Making Mixer Measurements

Where to Look for More Information 3-1

Measurement Considerations 3-2

Minimizing Source and Load Mismatches 3-2

Reducing the E ect of Spurious Responses 3-2

Eliminating Unwanted Mixing and Leakage Signals 3-2

How RF and IF Are De ned 3-2

Frequency O set Mode Operation 3-4

Di erences Between Internal and External R-Channel Inputs 3-4

P ower Meter Calibration 3-6

Conversion Loss Using the Frequency O set Mode 3-7

High Dynamic Range Swept RF/IF Conversion Loss 3-12

Fixed IF Mixer Measurements 3-17

Tuned Receiver Mode 3-17

Sequence 1 Setup 3-17

Sequence 2 Setup 3-21

Phase or Group Delay Measurements 3-24

Amplitude and Phase Tracking 3-27

Conversion Compression Using the Frequency O set Mode 3-28

Isolation Example Measurements 3-33

LO to RF Isolation 3-33

RF F eedthrough 3-35

Contents-3

Page 15 of the Agilent Technologies 8753D user manual — Table of Contents
Page 15 · click to enlarge
Page 15
4. Printing, Plotting, and Saving Measurement Results

Where to Look for More Information 4-2

Printing or Plotting Y our Measurement Results 4-3

Con guring a Print Function 4-3

De ning a Print Function 4-5

If Y ou are Using a Color Printer 4-6

T o Reset the Printing P arameters to Default V alues 4-6

Printing One Measurement P er P age 4-6

Printing Multiple Measurements P er P age 4-7

Con guring a Plot Function 4-8

If Y ou are Plotting to an HPGL/2 Compatible Printer 4-8

If Y ou are Plotting to a P en Plotter 4-10

If Y ou are Plotting to a Disk Drive 4-11

De ning a Plot Function 4-12

Choosing Display Elements 4-12

Selecting A uto-F eed 4-12

Selecting P en Numbers and Colors 4-13

Selecting Line Types 4-14

Choosing Scale 4-15

Choosing Plot Speed 4-15

T o Reset the Plotting P arameters to Default V alues 4-16

Plotting One Measurement P er P age Using a P en Plotter 4-16

Plotting Multiple Measurements P er P age Using a P en Plotter 4-17

If Y ou are Plotting to an HPGL Compatible Printer 4-18

Plotting a Measurement to Disk 4-19

T o Output the Plot Files 4-20

T o View Plot Files on a PC 4-20

Using AmiPro 4-21

Using Freelance 4-22

Outputting Plot Files from a PC to a Plotter 4-22

Outputting Plot Files from a PC to an HPGL Compatible Printer 4-23

1. Store the HPGL initialization sequence 4-23
2. Store the exit HPGL mode and form feed sequence 4-24
3. Send the HPGL initialization sequence to the printer 4-24
4. Send the plot le to the printer 4-24
5. Send the exit HPGL mode and form feed sequence to the printer 4-24

Outputting Single P age Plots Using a Printer 4-24

Outputting Multiple Plots to a Single P age Using a Printer 4-25

Plotting Multiple Measurements P er P age From Disk 4-26

T o Plot Multiple Measurements on a Full P age 4-26

T o Plot Measurements in P age Quadrants 4-28

Titling the Displayed Measurement 4-29

Con guring the Analyzer to Produce a Time Stamp 4-30

Aborting a Print or Plot Process 4-30

Printing or Plotting the List V alues or Operating P arameters 4-30

If Y ou want a Single P age of V alues 4-30

If Y ou W ant the Entire List of V alues 4-31

Solving Problems with Printing or Plotting 4-32

Saving and Recalling Instrument States 4-33

Places Where Y ou Can Save 4-33

What Y ou Can Save to the Analyzer's Internal Memory 4-33

What Y ou Can Save to a Floppy Disk 4-33

What Y ou Can Save to a Computer 4-34

Saving an Instrument State 4-35

Saving Measurement Results 4-36

Contents-4

Page 16 of the Agilent Technologies 8753D user manual — Table of Contents
Page 16 · click to enlarge
Page 16

ASCII Data F ormats 4-39

CITI le 4-39

S2P Data F ormat 4-39

Re-Saving an Instrument State 4-41

Deleting a File 4-41

T o Delete an Instrument State File 4-41

T o Delete all Files 4-41

Renaming a File 4-42

Recalling a File 4-42

F ormatting a Disk 4-43

Solving Problems with Saving or Recalling Files 4-43

If Y ou are Using an External Disk Drive 4-43

5. Optimizing Measurement Results

Where to Look for More Information 5-2

Increasing Measurement A ccuracy 5-2

Connector Repeatability 5-2

Interconnecting Cables 5-2

T emperature Drift 5-2

Frequency Drift 5-3

P erformance V eri cation 5-3

Reference Plane and P ort Extensions 5-3

Measurement Error-Correction 5-4

Conditions Where Error-Correction is Suggested 5-4

Types of Error-Correction 5-4

Error-Correction Stimulus State 5-5

Calibration Standards 5-6

Compensating for the Electrical Delay of Calibration Standards 5-6

Clarifying Type-N Connector Sex 5-6

When to Use Interpolated Error-Correction 5-6

Procedures for Error-Correcting Y our Measurements 5-7

Frequency Response Error-Corrections 5-8

Response Error-Correction for Re ection Measurements 5-8

Response Error-Correction for Transmission Measurements 5-10

Receiver Calibration 5-11

Frequency Response and Isolation Error-Corrections 5-13

Response and Isolation Error-Correction for Re ection Measurements 5-13

Response and Isolation Error-Correction for Transmission Measurements . . . 5-15

One-P ort Re ection Error-Correction 5-17

Full Two-P ort Error-Correction 5-20

TRL* and TRM* Error-Correction 5-23

TRL Error-Correction 5-23

TRM Error-Correction 5-24

Modifying Calibration Kit Standards 5-26

De nitions 5-26

Outline of Standard Modi cation 5-26

Modifying Standards 5-26

Modifying TRL Standards 5-28

Modifying TRM Standards 5-30

P ower Meter Measurement Calibration 5-33

Entering the P ower Sensor Calibration Data 5-34

Editing Frequency Segments 5-34

Deleting Frequency Segments 5-35

Compensating for Directional Coupler Response 5-35

Using Sample-and-Sweep Correction Mode 5-36

Contents-5

Page 17 of the Agilent Technologies 8753D user manual — Table of Contents
Page 17 · click to enlarge
Page 17

Using Continuous Correction Mode 5-37

T o Calibrate the Analyzer Receiver to Measure Absolute P ower 5-38

Matched A dapters 5-39

Modify the Cal Kit Thru De nition 5-40

Calibrating for Noninsertable Devices 5-41

A dapter Removal 5-42

P erform the 2-port Error Corrections 5-43

Remove the A dapter 5-44

V erify the Results 5-45

Example Program 5-47

Making A ccurate Measurements of Electrically Long Devices 5-48

The Cause of Measurement Problems 5-48

T o Improve Measurement Results 5-48

Decreasing the Sweep Rate 5-48

Decreasing the Time Delay 5-49

Increasing Sweep Speed 5-50

T o Decrease the Frequency Span 5-50

T o Set the A uto Sweep Time Mode 5-51

T o Widen the System Bandwidth 5-52

T o Reduce the A veraging F actor 5-52

T o Reduce the Number of Measurement P oints 5-52

T o Set the Sweep Type 5-53

T o View a Single Measurement Channel 5-53

T o A ctivate Chop Sweep Mode 5-54

T o Use External Calibration 5-54

T o Use F ast 2-P ort Calibration 5-54

Increasing Dynamic Range 5-56

T o Increase the T est P ort Input P ower 5-56

T o Reduce the Receiver Noise Floor 5-56

Changing System Bandwidth 5-56

Changing Measurement A veraging 5-56

Reducing Trace Noise 5-57

T o A ctivate A veraging 5-57

T o Change System Bandwidth 5-57

Reducing Receiver Crosstalk 5-57

Reducing Recall Time 5-58

Understanding Spur A voidance 5-59

6. Application and Operation Concepts

Where to Look for More Information 6-1

HP 8753D System Operation 6-2

The Built-In Synthesized Source 6-2

The Source Step Attenuator 6-2

The Built-In T est Set 6-3

The Receiver Block 6-3

The Microprocessor 6-3

Required P eripheral Equipment 6-3

Data Processing 6-4

Processing Details 6-5

The ADC 6-5

IF Detection 6-5

Ratio Calculations 6-5

Sampler/IF Correction 6-5

Sweep-T o-Sweep A veraging 6-5

Pre-Raw Data Arrays 6-6

Contents-6

Page 18 of the Agilent Technologies 8753D user manual — ENTRY OFF
Page 18 · click to enlarge
Page 18

Raw Arrays 6-6

V ector Error-correction (A ccuracy Enhancement) 6-6

Trace Math Operation 6-6

Gating (Option 010 Only) 6-6

The Electrical Delay Block 6-6

Conversion 6-6

Transform (Option 010 Only) 6-6

F ormat 6-7

Smoothing 6-7

F ormat Arrays 6-7

O set and Scale 6-7

Display Memory 6-7

A ctive Channel Keys 6-8

Dual Channel 6-8

Uncoupling Stimulus V alues Between Channels 6-8

Coupled Markers 6-8

Entry Block Keys 6-9

Units T erminator 6-9

Knob 6-10

Step Keys 6-10

ENTRY OFF

6-10

6-10

6-10

6-10

Stimulus Functions 6-11

De ning Ranges with Stimulus Keys 6-11

Stimulus Menu 6-12

The P ower Menu 6-13

Understanding the P ower Ranges 6-13

A utomatic mode 6-13

Manual mode 6-13

P ower Coupling Options 6-15

Channel coupling 6-15

T est port coupling 6-15

Sweep Time 6-16

Manual Sweep Time Mode 6-16

A uto Sweep Time Mode 6-16

Minimum Sweep Time 6-16

Trigger Menu 6-18

Source Attenuator Switch Protection 6-19

Allowing Repetitive Switching of the Attenuator 6-19

Channel Stimulus Coupling 6-20

Sweep Type Menu 6-21

Linear Frequency Sweep (Hz) 6-21

Logarithmic Frequency Sweep (Hz) 6-22

List Frequency Sweep (Hz) 6-22

Segment Menu 6-22

P ower Sweep (dBm) 6-23

CW Time Sweep (Seconds) 6-23

Selecting Sweep Modes 6-23

Modifying List Frequencies 6-23

Edit list menu 6-23

Edit subsweep menu 6-24

Response Functions 6-25

S-P arameters 6-26

Show 7 more labels / short text on this page
· · · · · · Contents-7
Page 19 of the Agilent Technologies 8753D user manual — Table of Contents
Page 19 · click to enlarge
Page 19

Understanding S-P arameters 6-26

The S-P arameter Menu 6-27

Analog In Menu 6-27

Conversion Menu 6-27

Input P orts Menu 6-28

The F ormat Menu 6-29

Log Magnitude F ormat 6-29

Phase F ormat 6-30

Group Delay F ormat 6-30

Smith Chart F ormat 6-31

P olar F ormat 6-32

Linear Magnitude F ormat 6-33

SWR F ormat 6-33

Real F ormat 6-34

Imaginary F ormat 6-34

Group Delay Principles 6-35

Scale Reference Menu 6-38

Electrical Delay 6-38

Display Menu 6-39

Dual Channel Mode 6-40

Dual Channel Mode with Decoupled Channel P ower 6-40

Memory Math Functions 6-41

A djusting the Colors of the Display 6-41

Setting Display Intensity 6-41

Setting Default Colors 6-42

Blanking the Display 6-42

Saving Modi ed Colors 6-42

Recalling Modi ed Colors 6-42

The Modify Colors Menu 6-42

A veraging Menu 6-44

A veraging 6-44

Smoothing 6-45

IF Bandwidth Reduction 6-45

Markers 6-47

Marker Menu 6-48

Delta Mode Menu 6-48

Fixed Marker Menu 6-48

Marker Function Menu 6-49

Marker Search Menu 6-49

T arget Menu 6-49

Marker Mode Menu 6-49

P olar Marker Menu 6-49

Smith Marker Menu 6-49

Measurement Calibration 6-50

What is A ccuracy Enhancement? 6-50

What Causes Measurement Errors? 6-51

Directivity 6-51

Source Match 6-52

Load Match 6-52

Isolation (Crosstalk) 6-53

Frequency Response (Tracking) 6-53

Characterizing Microwave Systematic Errors 6-53

One-P ort Error Model 6-53

Device Measurement 6-59

Two-P ort Error Model 6-59

Contents-8

Page 20 of the Agilent Technologies 8753D user manual — S
Page 20 · click to enlarge
Page 20

Calibration Considerations 6-65

Measurement P arameters 6-65

Device Measurements 6-65

Omitting Isolation Calibration 6-65

Saving Calibration Data 6-65

The Calibration Standards 6-66

Frequency Response of Calibration Standards 6-66

Electrical O set 6-67

Fringe Capacitance 6-67

How E ective Is A ccuracy Enhancement? 6-69

Correcting for Measurement Errors 6-71

Ensuring a V alid Calibration 6-71

Interpolated Error-correction 6-72

The Calibrate Menu 6-73

Response Calibration 6-73

Response and Isolation Calibration 6-73

S

One-P ort Calibration 6-73

Full Two-P ort Calibration 6-73

TRL*/LRM* Two-P ort Calibration 6-74

Restarting a Calibration 6-75

Cal Kit Menu 6-75

The Select Cal Kit Menu 6-75

Modifying Calibration Kits 6-76

De nitions 6-76

Procedure 6-76

Modify Calibration Kit Menu 6-77

De ne Standard Menus 6-78

Specify O set Menu 6-80

Label Standard Menu 6-81

Specify Class Menu 6-81

Label Class Menu 6-83

Label Kit Menu 6-83

V erify performance 6-84

TRL*/LRM* Calibration 6-85

Why Use TRL Calibration? 6-85

TRL T erminology 6-85

How TRL*/LRM* Calibration W orks 6-86

TRL* Error Model 6-86

Isolation 6-87

Source match and load match 6-88

Improving Raw Source Match and Load Match F or TRL*/LRM* Calibration . . 6-88

The TRL Calibration Procedure 6-90

Requirements for TRL Standards 6-90

F abricating and de ning calibration standards for TRL/LRM 6-91

TRL Options 6-93

P ower Meter Calibration 6-95

Primary Applications 6-95

Calibrated P ower Level 6-95

Compatible Sweep Types 6-95

Loss of P ower Meter Calibration Data 6-96

Interpolation in P ower Meter Calibration 6-96

P ower Meter Calibration Modes of Operation 6-96

Continuous Sample Mode (Each Sweep) 6-96

Sample-and-Sweep Mode (One Sweep) 6-97

P ower Loss Correction List 6-98

and S · Contents-9

Page 21 of the Agilent Technologies 8753D user manual — LOCAL
Page 21 · click to enlarge
Page 21

P ower Sensor Calibration F actor List 6-98

Speed and A ccuracy 6-98

T est Equipment Used 6-98

Stimulus P arameters 6-98

Alternate and Chop Sweep Modes 6-100

Alternate 6-100

Chop 6-100

Calibrating for Non-Insertable Devices 6-101

A dapter Removal 6-101

Matched A dapters 6-101

Modify the Cal Kit Thru De nition 6-101

Using the Instrument State Functions 6-102

HP-IB Menu 6-103

LOCAL

Key 6-103

HP-IB ST A TUS Indicators 6-104

System Controller Mode 6-104

T alker/Listener Mode 6-104

P ass Control Mode 6-104

A ddress Menu 6-104

Using the P arallel P ort 6-105

The Copy Mode 6-105

The GPIO Mode 6-105

The System Menu 6-106

The Limits Menu 6-106

Edit Limits Menu 6-107

Edit Segment Menu 6-107

O set Limits Menu 6-108

Knowing the Instrument Modes 6-109

Network Analyzer Mode 6-109

External Source Mode 6-109

Primary Applications 6-109

Typical T est Setup 6-110

External Source Mode In-Depth Description 6-110

External Source A uto 6-110

External Source Manual 6-110

CW Frequency Range in External Source Mode 6-111

Compatible Sweep Types 6-111

External Source Requirements 6-111

Capture Range 6-111

Locking onto a signal with a frequency modulation component 6-111

Tuned Receiver Mode 6-111

Frequency O set Menu 6-112

Primary Applications 6-112

Typical T est Setup 6-112

Frequency O set In-Depth Description 6-113

The Receiver Frequency 6-113

The O set Frequency (LO) 6-113

Frequency Hierarchy 6-113

Frequency Ranges 6-113

Compatible Instrument Modes and Sweep Types 6-113

Receiver and Source Requirements 6-114

Display Annotations 6-114

Error Message 6-114

Spurious Signal P assband Frequencies 6-114

· Contents-10

Page 22 of the Agilent Technologies 8753D user manual — Table of Contents
Page 22 · click to enlarge
Page 22

Harmonic Operation (Option 002 only) 6-115

Typical T est Setup 6-115

Single-Channel Operation 6-115

Dual-Channel Operation 6-115

Coupling P ower Between Channels 1 and 2 6-115

Frequency Range 6-116

A ccuracy and input power 6-116

Time Domain Operation (Option 010) 6-117

The Transform Menu 6-117

General Theory 6-118

Time Domain Bandpass 6-119

A djusting the Relative V elocity F actor 6-119

Re ection Measurements Using Bandpass Mode 6-119

Interpreting the bandpass re ection response horizontal axis 6-120

Interpreting the bandpass re ection response vertical axis 6-120

Transmission Measurements Using Bandpass Mode 6-121

Interpreting the bandpass transmission response horizontal axis 6-121

Interpreting the bandpass transmission response vertical axis 6-121

Time domain low pass 6-122

Setting frequency range for time domain low pass 6-122

Minimum allowable stop frequencies 6-123

Re ection Measurements In Time Domain Low P ass 6-123

Interpreting the low pass response horizontal axis 6-123

Interpreting the low pass response vertical axis 6-123

F ault Location Measurements Using Low P ass 6-123

Transmission Measurements In Time Domain Low P ass 6-125

Measuring small signal transient response using low pass step 6-125

Interpreting the low pass step transmission response horizontal axis . . . 6-126

Interpreting the low pass step transmission response vertical axis 6-126

Measuring separate transmission paths through the test device using low

pass impulse mode 6-126

Time Domain Concepts 6-127

Masking 6-127

Windowing 6-128

Range 6-130

Resolution 6-131

Response resolution 6-131

Range resolution 6-132

Gating 6-133

Setting the gate 6-133

Selecting gate shape 6-134

Transforming CW Time Measurements Into the Frequency Domain 6-134

F orward Transform Measurements 6-135

Interpreting the forward transform vertical axis 6-135

Interpreting the forward transform horizontal axis 6-135

Demodulating the results of the forward transform 6-135

F orward transform range 6-136

T est Sequencing 6-138

In-Depth Sequencing Information 6-138

F eatures That Operate Di erently When Executed In a Sequence 6-138

Commands That Sequencing Completes Before the Next Sequence Command

Begins 6-138

Commands That Require a Clean Sweep 6-139

F orward Stepping In Edit Mode 6-139

Titles 6-139

Contents-11

Page 23 of the Agilent Technologies 8753D user manual — Table of Contents
Page 23 · click to enlarge
Page 23

Sequence Size 6-139

Embedding the V alue of the Loop Counter In a Title 6-139

A utostarting Sequences 6-139

The GPIO Mode 6-139

The Sequencing Menu 6-140

Gosub Sequence Command 6-140

TTL I/O Menu 6-140

TTL Output for Controlling P eripherals 6-140

TTL Input Decision Making 6-140

TTL Out Menu 6-142

Sequencing Special Functions Menu 6-142

Sequence Decision Making Menu 6-142

Decision Making Functions 6-142

Decision making functions jump to a softkey location, not to a speci c

sequence title 6-142

Having a sequence jump to itself 6-142

TTL input decision making 6-142

Limit test decision making 6-142

Loop counter decision making 6-143

Naming Files Generated by a Sequence 6-143

HP-GL Considerations 6-143

Entering HP-GL Commands 6-143

Special Commands 6-144

Entering Sequences Using HP-IB 6-144

Reading Sequences Using HP-IB 6-144

Ampli er T esting 6-145

Ampli er parameters 6-145

Gain Compression 6-146

Metering the power level 6-148

Mixer T esting 6-149

Frequency O set 6-149

Tuned Receiver 6-149

Mixer P arameters That Y ou Can Measure 6-150

A ccuracy Considerations 6-150

Attenuation at Mixer P orts 6-151

Filtering 6-152

Frequency Selection 6-153

LO Frequency A ccuracy and Stability 6-153

Up-Conversion and Down-Conversion De nition 6-153

Conversion Loss 6-156

Isolation 6-156

LO F eedthru / LO to RF Leakage 6-156

RF F eedthru 6-157

SWR / Return Loss 6-157

Conversion Compression 6-158

Phase Measurements 6-158

Amplitude and Phase Tracking 6-159

Phase Linearity and Group Delay 6-159

Connection Considerations 6-161

A dapters 6-161

Fixtures 6-162

If Y ou W ant to Design Y our Own Fixture 6-162

Reference Documents 6-163

General Measurement and Calibration T echniques 6-163

Fixtures and Non-Coaxial Measurements 6-163

Contents-12

Page 24 of the Agilent Technologies 8753D user manual — Table of Contents
Page 24 · click to enlarge
Page 24

On-W afer Measurements 6-164

7. Speci cations and Measurement Uncertainties

Dynamic Range 7-1

HP 8753D Network Analyzer Speci cations 7-2

HP 8753D (50 ) with 7 mm T est P orts 7-2

Measurement P ort Characteristics 7-2

Transmission Measurement Uncertainties 7-3

Re ection Measurement Uncertainties 7-4

HP 8753D (50 ) with Type-N T est P orts 7-5

Measurement P ort Characteristics 7-5

Transmission Measurement Uncertainties 7-6

Re ection Measurement Uncertainties 7-7

HP 8753D (50 ) with 3.5 mm T est P orts 7-8

Measurement P ort Characteristics 7-8

Transmission Measurement Uncertainties 7-9

Re ection Measurement Uncertainties 7-10

HP 8753D (75 ) with Type-N T est P orts 7-11

Measurement P ort Characteristics 7-11

Transmission Measurement Uncertainties 7-12

Re ection Measurement Uncertainties 7-13

HP 8753D (75 ) with Type-F T est P orts 7-14

Measurement P ort Characteristics 7-14

Transmission Measurement Uncertainties 7-15

Re ection Measurement Uncertainties 7-16

Transmission Measurement Uncertainties 7-18

Re ection Measurement Uncertainties 7-19

Transmission Measurement Uncertainties 7-21

Re ection Measurement Uncertainties 7-22

Transmission Measurement Uncertainties 7-24

Re ection Measurement Uncertainties 7-25

Instrument Speci cations 7-26

HP 8753D Network Analyzer General Characteristics 7-33

Measurement Throughput Summary 7-33

Remote Programming 7-33

Interface 7-33

Transfer F ormats 7-33

Interface Function Codes 7-33

Front P anel Connectors 7-34

Probe P ower 7-34

Rear P anel Connectors 7-34

External Reference Frequency Input (EXT REF INPUT) 7-34

High-Stability Frequency Reference Output (10 MHz)(Option 001) 7-34

External A uxiliary Input (A UX INPUT) 7-34

External AM Input (EXT AM) 7-34

External Trigger (EXT TRIGGER) 7-35

T est Sequence Output (TEST SEQ) 7-35

Limit T est Output (LIMIT TEST) 7-35

T est P ort Bias Input (BIAS CONNECT) 7-35

Video Output (EXT MON) 7-35

HP-IB 7-35

P arallel P ort 7-36

RS-232 7-36

DIN Keyboard 7-36

Line P ower 7-36

Contents-13

Page 25 of the Agilent Technologies 8753D user manual — Table of Contents
Page 25 · click to enlarge
Page 25

Environmental Characteristics 7-36

General Conditions 7-36

Operating Conditions 7-36

Non-Operating Storage Conditions 7-36

W eight 7-37

Cabinet Dimensions 7-37

Internal Memory 7-37

8. Menu Maps
9. K ey De nitions

Where to Look for More Information 9-1

Guide T erms and Conventions 9-2

Analyzer Functions 9-2

Cross Reference of Key Function to Programming Command 9-48

Softkey Locations 9-69

10. Error Messages

Where to Look for More Information 10-1

Error Messages in Alphabetical Order 10-2

Error Messages in Numerical Order 10-26

11. Compatible P eripherals

Where to Look for More Information 11-1

Measurement A ccessories A vailable 11-1

Calibration Kits 11-1

V eri cation Kit 11-2

HP 85029B 7 mm V eri cation Kit 11-2

T est P ort Return Cables 11-2

HP 11857D 7 mm T est P ort Return Cable Set 11-2

HP 11857B 75 Ohm Type-N T est P ort Return Cable Set 11-2

A dapter Kits 11-2

HP 11852B 50 to 75 Ohm Minimum Loss P ad 11-2

Transistor T est Fixtures 11-3

HP 11600B and 11602B Transistor Fixtures 11-3

HP 11608A Option 003 Transistor Fixture 11-3

HP 11858A Transistor Fixture A dapter 11-3

System A ccessories A vailable 11-4

System Cabinet 11-4

System T estmobile 11-4

Plotters and Printers 11-4

These plotters are compatible: 11-4

These printers are compatible: 11-4

Mass Storage 11-5

HP-IB Cables 11-5

Interface Cables 11-5

Keyboards 11-6

Controller 11-6

Sample Software 11-6

External Monitors 11-6

Connecting P eripherals 11-7

Connecting the P eripheral Device 11-7

Con guring the Analyzer for the P eripheral 11-8

If the P eripheral is a Printer 11-8

If the P eripheral is a Plotter 11-9

Contents-14

Page 26 of the Agilent Technologies 8753D user manual — Table of Contents
Page 26 · click to enlarge
Page 26

HPGL/2 Compatible Printer (used as a plotter) 11-9

P en Plotter 11-11

If the P eripheral is a P ower Meter 11-12

If the P eripheral is an External Disk Drive 11-12

If the P eripheral is a Computer Controller 11-12

Con guring the Analyzer to Produce a Time Stamp 11-13

HP-IB Programming Overview 11-14

HP-IB Operation 11-15

Device Types 11-15

T alker 11-15

Listener 11-15

Controller 11-15

HP-IB Bus Structure 11-16

Data Bus 11-16

Handshake Lines 11-16

Control Lines 11-16

HP-IB Requirements 11-17

HP-IB Operational Capabilities 11-18

HP-IB Status Indicators 11-19

Bus Device Modes 11-19

System-Controller Mode 11-20

T alker/Listener Mode 11-20

P ass-Control Mode 11-20

Setting HP-IB A ddresses 11-20

Analyzer Command Syntax 11-21

Code Naming Convention 11-21

V alid Characters 11-22

Units 11-23

HP-IB Debug Mode 11-23

12. Preset State and Memory Allocation

Where to Look for More Information 12-1

Types of Memory and Data Storage 12-1

V olatile Memory 12-1

Non-V olatile Memory 12-2

Storing Data to Disk 12-4

Conserving Memory 12-5

Using Saved Calibration Sets 12-5

Preset State 12-6

A. The CITI le Data F ormat and K eyword Reference

The CITI le Data F ormat A -1

Description and Overview A -1

Data F ormats A -1

File and Operating System F ormats A -1

De nition of CITI le T erms A -2

A CITI le P ackage A -2

The CITI le Header A -2

An Array of Data A -2

CITI le Keyword A -3

CITI le Examples A -4

Example 2, An 8510 Display Memory File A -4

Example 3, 8510 Data le A -4

Example 4, 8510 3-T erm Frequency List Cal Set File A -5

Conclusion A -6

Contents-15

Page 27 of the Agilent Technologies 8753D user manual — Table of Contents
Page 27 · click to enlarge
Page 27

The CITI le Keyword Reference A -7

Index · Contents-16

Page 28 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 28 · click to enlarge
Page 28

1-1. HP 8753D Front P anel 1-4

1-2. Analyzer Display (Single Channel, Cartesian F ormat) 1-6

1-3. HP 8753D Rear P anel 1-10

2-1. Basic Measurement Setup 2-3

2-2. Example of Viewing Both Channels Simultaneously 2-5

2-3. Example Dual Channel With Split Display On 2-6

2-4. Example of a Display Title 2-8

2-5. A ctive Marker Control Example 2-10

2-6. A ctive and Inactive Markers Example 2-10

2-7. Marker 1 as the Reference Marker Example 2-11

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

2-8. Example of a Fixed Reference Marker Using

REF=FIXED MKR 2-12

NNNNNNNNNNNNNNNNNNNNNNNNNN

2-9. Example of a Fixed Reference Marker Using

MKR ZERO 2-13

2-10. Example of Coupled and Uncoupled Markers 2-14

2-11. Example of a Log Marker in P olar F ormat 2-15

2-12. Example of Impedance Smith Chart Markers 2-16

2-13. Example of Setting the Start Frequency Using a Marker 2-17

2-14. Example of Setting the Stop Frequency Using a Marker 2-17

2-15. Example of Setting the Center Frequency Using a Marker 2-18

2-16. Example of Setting the Frequency Span Using Markerj 2-19

2-17. Example of Setting the Reference V alue Using a Marker 2-20

2-18. Example of Setting the Electrical Delay Using a Marker 2-21

2-19. Example of Searching for the Maximum Amplitude Using a Marker 2-22

2-20. Example of Searching for the Minimum Amplitude Using a Marker 2-23

2-21. Example of Searching for a T arget Amplitude Using a Marker 2-24

2-22. Example of Searching for a Bandwidth Using Markers 2-25

2-23. Example Statistics of Measurement Data 2-26

2-24. Device Connections for Measuring a Magnitude Response 2-27

2-25. Example Magnitude Response Measurement Results 2-28

2-26. Example Insertion Phase Response Measurement 2-28

2-27. Phase Samples 2-29

2-28. Device Connections for Measuring Electrical Length 2-30

2-29. Linearly Changing Phase 2-31

2-30. Example Best Flat Line with A dded Electrical Delay 2-32

2-31. Deviation From Linear Phase Example Measurement 2-33

2-32. Group Delay Example Measurement 2-34

2-33. Group Delay Example Measurement with Smoothing 2-34

2-34. Group Delay Example Measurement with Smoothing Aperture Increased . . . 2-35

2-35. Connections for SA W Filter Example Measurement 2-36

2-36. Example Flat Limit Line 2-38

2-37. Example Flat Limit Lines 2-39

2-38. Sloping Limit Lines 2-40

2-39. Example Single P oints Limit Line 2-41

2-40. Example Stimulus O set of Limit Lines 2-44

2-41. Diagram of Gain Compression 2-45

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

2-42. Gain Compression Using Linear Sweep and

D2/D1 to D2 ON 2-46

Figures · Contents-17

Page 29 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 29 · click to enlarge
Page 29

2-43. Gain Compression Using P ower Sweep 2-48

2-44. Gain and Reverse Isolation 2-50

2-45. Typical T est Setup for Tuned Receiver Mode 2-51

2-46. T est Sequencing Help Instructions 2-54

2-47. Fundamental and 2nd Harmonic P ower Levels in dBm 2-66

2-48. 2nd Harmonic P ower Level in dBc 2-67

2-49. Device Connections for Time Domain Transmission Example Measurement . . 2-68

2-50. Time Domain Transmission Example Measurement 2-69

2-51. Gating in a Time Domain Transmission Example Measurement 2-70

2-52. Gate Shape 2-71

2-53. Gating E ects in a Frequency Domain Example Measurement 2-72

2-54. Device Connections for Re ection Time Domain Example Measurement 2-73

2-55. Device Response in the Frequency Domain 2-74

2-56. Device Response in the Time Domain 2-75

3-1. Down Converter P ort Connections 3-3

3-2. Up Converter P ort Connections 3-3

3-3. R-Channel External Connection 3-4

3-4. An Example Spectrum of RF , LO , and IF Signals Present in a Conversion Loss

Measurement 3-7

3-5. Connections for R Channel and Source Calibration 3-8

3-6. Connections for a One-Sweep P ower Meter Calibration for Mixer Measurements 3-9

3-7. Diagram of Measurement Frequencies 3-10

3-8. Measurement Setup from Display 3-10

3-9. Conversion Loss Example Measurement 3-11

3-10. Connections for Broad Band P ower Meter Calibration 3-13

3-11. Connections for Receiver Calibration 3-14

3-12. Connections for a High Dynamic Range Swept IF Conversion Loss Measurement 3-15

3-13. Example of Swept IF Conversion Loss Measurement 3-16

3-14. Connections for a Response Calibration 3-18

3-15. Connections for a Conversion Loss Using the Tuned Receiver Mode 3-22

3-16. Example Fixed IF Mixer Measurement 3-23

3-17. Connections for a Group Delay Measurement 3-25

3-18. Group Delay Measurement Example 3-26

3-19. Conversion Loss and Output P ower as a Function of Input P ower Level Example 3-28

3-20. Connections for the First P ortion of Conversion Compression Measurement . . 3-29

3-21. Connections for the Second P ortion of Conversion Compression Measurement . 3-30

3-22. Measurement Setup Diagram Shown on Analyzer Display 3-31

3-23. Example Swept P ower Conversion Compression Measurement 3-32

3-24. Signal Flow in a Mixer Example 3-33

3-25. Connections for a Response Calibration 3-34

3-26. Connections for a Mixer Isolation Measurement 3-34

3-27. Example Mixer LO to RF Isolation Measurement 3-35

3-28. Connections for a Response Calibration 3-36

3-29. Connections for a Mixer RF F eedthrough Measurement 3-36

3-30. Example Mixer RF F eedthrough Measurement 3-37

4-1. Printer Connections to the Analyzer 4-3

4-2. Printing Two Measurements 4-7

4-3. P eripheral Connections to the Analyzer 4-8

4-4. Plot Components A vailable through De nition 4-12

4-5. Line Types A vailable 4-14

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

4-6. Locations of P1 and P2 in

SCALE PLOT [GRAT] Mode 4-15

4-7. Plot Quadrants 4-17

4-8. A utomatic File Naming Convention for LIF F ormat 4-19

4-9. Plot Filename Convention 4-26

4-10. Plotting Two Files on the Same P age 4-27

Contents-18

Page 30 of the Agilent Technologies 8753D user manual — SF
Page 30 · click to enlarge
Page 30

4-11. Plot Quadrants 4-28

4-12. Data Processing Flow Diagram 4-37

5-1. Standard Connections for a Response Error-Correction for Re ection

Measurement 5-9

5-2. Standard Connections for Response Error-Correction for Transmission

Measurements 5-10

5-3. Standard Connections for Receiver Calibration 5-11

5-4. Standard Connections for a Response and Isolation Error-Correction for

Re ection Measurements 5-14

5-5. Standard Connections for a Response and Isolation Error-Correction for

Transmission Measurements 5-16

5-6. Standard Connections for a One P ort Re ection Error-Correction 5-18

5-7. Standard Connections for Full Two port Error-Correction 5-20

5-8. Sample-and-Sweep Mode for P ower Meter Calibration 5-36

5-9. Continuous Correction Mode for P ower Meter Calibration 5-37

5-10. Calibrating for Noninsertable Devices 5-39

5-11. Noninsertable Device 5-41

5-12. A dapters Needed 5-42

5-13. Two-P ort Cal Set 1 5-43

5-14. Two-P ort Cal Set 2 5-44

5-15. Calibrated Measurement 5-45

6-1. Simpli ed Block Diagram of the Network Analyzer System 6-2

6-2. Data Processing Flow Diagram 6-4

6-3. A ctive Channel Keys 6-8

6-4. Entry Block 6-9

6-5. Stimulus Function Block 6-11

6-6. P ower Range Transitions in the A utomatic Mode 6-14

6-7. Response Function Block 6-25

6-8. S-P arameters of a Two-P ort Device 6-26

6-9. Re ection Impedance and A dmittance Conversions 6-28

6-10. Transmission Impedance and A dmittance Conversions 6-28

6-11. Log Magnitude F ormat 6-30

6-12. Phase F ormat 6-30

6-13. Group Delay F ormat 6-31

6-14. Standard and Inverse Smith Chart F ormats 6-32

6-15. P olar F ormat 6-32

6-16. Linear Magnitude F ormat 6-33

6-17. Typical SWR Display 6-33

6-18. Real F ormat 6-34

6-19. Constant Group Delay 6-35

6-20. Higher Order Phase Shift 6-35

6-21. Rate of Phase Change V ersus Frequency 6-36

6-22. V ariations in Frequency Aperture 6-36

6-23. Dual Channel Displays 6-40

6-24. E ect of A veraging on a Trace 6-44

6-25. E ect of Smoothing on a Trace 6-45

6-26. IF Bandwidth Reduction 6-46

6-27. Markers on Trace 6-47

6-28. Directivity 6-51

6-29. Source Match 6-52

6-30. Load Match 6-52

6-31. Sources of Error in a Re ection Measurement 6-54

6-32. Re ection Coecient 6-54

6-33. E ective Directivity E

6-55

DF

6-34. Source Match E

6-55

SF

Contents-19

Page 31 of the Agilent Technologies 8753D user manual — XF
Page 31 · click to enlarge
Page 31

6-35. Re ection Tracking E

6-56

RF

6-36. \P erfect Load" T ermination 6-56

6-37. Measured E ective Directivity 6-57

6-38. Short Circuit T ermination 6-57

6-39. Open Circuit T ermination 6-58

6-40. Measured S

6-59

6-41. Major Sources of Error 6-60

6-42. Transmission Coecient 6-60

6-43. Load Match E

LF

6-61

6-44. Isolation E

XF

6-62

6-45. Full Two-P ort Error Model 6-63

6-46. Full Two-P ort Error Model Equations 6-64

6-47. Typical Responses of Calibration Standards after Calibration 6-68

6-48. Response versus S

1-P ort Calibration on Log Magnitude F ormat 6-69

6-49. Response versus S

1-P ort Calibration on Smith Chart 6-70

6-50. Response versus Full Two-P ort Calibration 6-70

6-51. HP 8753D functional block diagram for a 2-port error-corrected measurement

system 6-86

6-52. 8-term TRL error model and generalized coecients 6-87

6-53. Typical Measurement Set up 6-89

6-54. T est Setup for Continuous Sample Mode 6-97

6-55. T est Setup for Sample-and-Sweep Mode 6-97

6-56. Alternate and Chop Sweeps Overlaid 6-100

6-57. Instrument State Function Block 6-102

6-58. Typical Setup for the External Source Mode 6-110

6-59. Typical T est Setup for a Frequency O set Measurement 6-113

6-60. Typical Harmonic Mode T est Setup 6-115

6-61. Device Frequency Domain and Time Domain Re ection Responses 6-118

6-62. A Re ection Measurement of Two Cables 6-120

6-63. Transmission Measurement in Time Domain Bandpass Mode 6-121

6-64. Time Domain Low P ass Measurements of an Unterminated Cable 6-123

6-65. Simulated Low P ass Step and Impulse Response W aveforms (Real F ormat) . . . 6-124

6-66. Low P ass Step Measurements of Common Cable F aults (Real F ormat) 6-125

6-67. Time Domain Low P ass Measurement of an Ampli er Small Signal Transient

Response 6-126

6-68. Transmission Measurements Using Low P ass Impulse Mode 6-127

6-69. Masking Example 6-128

6-70. Impulse Width, Sidelobes , and Windowing 6-128

6-71. The E ects of Windowing on the Time Domain Responses of a Short Circuit . . 6-130

6-72. Response Resolution 6-132

6-73. Range Resolution of a Single Discontinuity 6-132

6-74. Sequence of Steps in Gating Operation 6-133

6-75. Gate Shape 6-134

6-76. Ampli er Gain Measurement 6-135

6-77. Combined E ects of Amplitude and Phase Modulation 6-136

6-78. Separating the Amplitude and Phase Components of T est-Device-Induced

Modulation 6-136

6-79. Range of a F orward Transform Measurement 6-137

6-80. P arallel P ort Input and Output Bus Pin Locations in GPIO Mode 6-141

6-81. Ampli er P arameters 6-145

6-82. Swept Frequency Ampli er Measurement of Absolute Fundamental, 2nd and 3rd

Harmonic Output Levels 6-145

6-83. Swept Frequency Ampli er Measurement of 2nd and 3rd Harmonic Distortion

(dBc) 6-146

6-84. Diagram of Gain Compression 6-147

Contents-20

Page 32 of the Agilent Technologies 8753D user manual — List of Figures
Page 32 · click to enlarge
Page 32

6-85. Swept P ower Measurement of Ampli er's Fundamental Gain Compression and

2nd Harmonic Output Level 6-147

6-86. T est Con guration for Setting RF Input using A utomatic P ower Meter Calibration 6-148

6-87. Mixer P arameters 6-150

6-88. Conversion Loss versus Output Frequency Without Attenuators at Mixer P orts 6-151

6-89. Example of Conversion Loss versus Output Frequency Without Correct IF Signal

P ath Filtering 6-152

6-90. Example of Conversion Loss versus Output Frequency With Correct IF Signal

P ath Filtering and Attenuation at all Mixer P orts 6-152

6-91. Examples of Up Converters and Down Converters 6-153

6-92. Down Converter P ort Connections 6-154

6-93. Up Converter P ort Connections 6-155

6-94. Example Spectrum of RF , LO , and IF Signals Present in a Conversion Loss

Measurement 6-156

6-95. Main Isolation T erms 6-156

6-96. Conversion Loss and Output P ower as a Function of Input P ower Level 6-158

6-97. Connections for an Amplitude and Phase Tracking Measurement Between Two

Mixers 6-159

6-98. A dapter Considerations 6-161

7-1. External Trigger Circuit 7-35

11-1. P eripheral Connections to the Analyzer 11-7

11-2. HP-IB Bus Structure 11-16

11-3. Analyzer Single Bus Concept 11-19

Contents-21

Page 33 of the Agilent Technologies 8753D user manual — List of Tables
Page 33 · click to enlarge
Page 33

0-1. Hewlett-P ackard Sales and Service Oces v

1-1. Comparing the HP 8753A/B/C/D 1-14

2-1. Connector Care Quick Reference 2-2

2-2. Gate Characteristics 2-71

4-1. Default V alues for Printing P arameters 4-6

4-2. Default P en Numbers and Corresponding Colors 4-13

4-3. Default P en Numbers for Plot Elements 4-13

4-4. Default Line Types for Plot Elements 4-14

4-5. Plotting P arameter Default V alues 4-16

4-6. HPGL Initialization Commands 4-23

4-7. HPGL T est File Commands 4-24

5-1. Di erences between PORT EXTENSIONS and ELECTRICAL DELA Y 5-3

5-2. Purpose and Use of Di erent Error-Correction Procedures 5-5

5-3. Typical Calibration Kit Standard and Corresponding Number 5-27

5-4. Characteristic P ower Meter Calibration Sweep Speed and A ccuracy 5-33

5-5. Band Switch P oints 5-50

5-6. Typical Recall State Times 5-58

6-1. Minimum Cycle Time (in seconds) 6-17

6-2. Display Colors with Maximum Viewing Angle 6-43

6-3. Calibration Standard Types and Expected Phase Shift 6-67

6-4. Standard De nitions 6-78

6-5. Standard Class Assignments 6-82

6-6. Characteristic P ower Meter Calibration Sweep Speed and A ccuracy 6-99

6-7. External Source Capture Ranges 6-111

6-8. Maximum Fundamental Frequency using Harmonic Mode 6-116

6-9. Time Domain Re ection F ormats 6-121

6-10. Minimum Frequency Ranges for Time Domain Low P ass 6-122

6-11. Impulse Width, Sidelobe Level, and Windowing V alues 6-129

6-12. Gate Characteristics 6-134

7-1. HP 8753D Dynamic Range 7-1

7-2. Measurement P ort Characteristics (Corrected

) for HP 8753D (50 ) with 7 mm

T est P orts 7-2

7-3. Measurement P ort Characteristics (Uncorrected

) for HP 8753D (50 ) with 7 mm

T est P orts 7-2

7-4. Measurement P ort Characteristics (Corrected)

for HP 8753D (50 ) with Type-N

T est P orts 7-5

7-5. Measurement P ort Characteristics (Corrected)

for HP 8753D (50 ) with 3.5 mm

T est P orts 7-8

7-6. Measurement P ort Characteristics (Corrected)

for HP 8753D (75 ) with Type-N

T est P orts 7-11

7-7. Measurement P ort Characteristics (Uncorrected)

for HP 8753D (75 ) with

Type-N T est P orts 7-11

7-8. Measurement P ort Characteristics (Corrected)

for HP 8753D (75 ) using HP

85039A F-M T est P orts 7-14

7-9. Measurement P ort Characteristics (Corrected)

for HP 8753D (75 ) using HP

85039A F-F T estports 7-17

Show 10 more labels / short text on this page
T ables · * · * · * · * · * · * y · * · * · Contents-22
Page 34 of the Agilent Technologies 8753D user manual — List of Tables
Page 34 · click to enlarge
Page 34

7-10. Measurement P ort Characteristics (Corrected)

for HP 8753D (75 ) using HP

85039A M-M T estports 7-20

7-11. Measurement P ort Characteristics (Corrected)

for HP 8753D (75 ) using HP

85039A M-F T estports 7-23

9-1. Cross Reference of Key Function to Programming Command 9-48

9-2. Softkey Locations 9-70

11-1. Code Naming Convention 11-22

12-1. Memory Requirements of Calibration and Memory Trace Arrays 12-3

12-2. Sux Character De nitions 12-4

12-3. Preset Conditions (1 of 5) 12-7

12-4. P ower-on Conditions (versus Preset) 12-10

12-5. Results of P ower Loss to Non-V olatile Memory 12-11

* · * · Contents-23

Page 35 of the Agilent Technologies 8753D user manual — List of Tables
Page 35 · click to enlarge
Page 35
Page 35 of the Agilent Technologies 8753D user manual — List of Tables
Page 36 of the Agilent Technologies 8753D user manual — Chapter 1 8753D Description and Options
Page 36 · click to enlarge
Page 36

HP 8753D Description and Options

This chapter contains information on the following topics:

Rear panel features and connectors

Service and support options

Changes between the HP 8753 network analyzers

Where to Look for More Information

A dditional information about many of the topics discussed in this chapter is located in the

following areas:

Chapter 2, \Making Measurements ," contains step-by-step procedures for making

measurements or using particular functions .

Chapter 4, \Printing, Plotting, and Saving Measurement Results ," contains instructions

for saving to disk or the analyzer internal memory , and printing and plotting displayed

measurements .

Chapter 5, \Optimizing Measurement Results ," describes techniques and functions for

achieving the best measurement results .

Chapter 6, \Application and Operation Concepts ," contains explanatory-style information

about many applications and analyzer operation.

HP 8753D Description and Options 1-1

Show 5 more labels / short text on this page
Analyzer overview · Analyzer description · Front panel features · Analyzer display · Analyzer options available
Page 37 of the Agilent Technologies 8753D user manual — Chapter 1 8753D Description and Options
Page 37 · click to enlarge
Page 37

The HP 8753D is a high performance vector network analyzer for laboratory or production

measurements of re ection and transmission parameters . It integrates a high resolution

synthesized RF source , an S-parameter test set, and a dual channel three-input receiver to

measure and display magnitude , phase , and group delay responses of active and passive RF

networks .

Two independent display channels and a large screen color display show the measured results

of one or both channels , in rectangular or polar/Smith chart formats .

F or information on options , refer to \Options A vailable" later in this chapter .

The analyzer has the additional following features:

Measurement functions selection with front panel keys and softkey menus .

External keyboard compatibility that allows you to title les and control the analyzer .

Internal automation, using test sequencing to program analyzer measurements and control

other devices without an external controller .

T est system automation with the addition of an external controller . This allows all of the

analyzer's measurement capabilities to be programmed over the Hewlett-P ackard Interface

Bus (HP-IB). (Refer to Chapter 11, \Compatible P eripherals"or the HP 8753D Network

Analyzer Programming Guide.)

A general purpose input/output (GPIO) bus that can control eight output bits and read ve

input bits through test sequencing. This can be useful for interfacing to material handlers

or custom test sets .

A utomatic sweep time that selects the minimum sweep time for the given IF bandwidth,

number of points , averaging mode , frequency range , and sweep type .

Built-in service diagnostics are available to simplify troubleshooting procedures .

Measurement exibility through trace math, data averaging, trace smoothing, electrical

delay , and accuracy enhancement.

External source mode capability that allows you to phase lock the analyzer's receiver to an

external source . Refer to Chapter 6, \Applications and Operation Concepts ."

Tuned receiver mode that allows you to use the receiver as a stand-alone device . Refer to

Chapter 6, \Applications and Operation Concepts ."

Complete re ection and transmission measurements in either 50 or 75 ohm impedance

environments .

Receiver/source frequency o set mode that allows you to set the analyzer's receiver and

source with a xed frequency o set for mixer test applications .

1-2 HP 8753D Description and Options

Analyzer Description · Control · P erformance

Page 38 of the Agilent Technologies 8753D user manual — Chapter 1 8753D Description and Options
Page 38 · click to enlarge
Page 38

A ccuracy enhancement methods that range from normalizing data to complete one or two

port vector error correction with up to 1601 measurement points , and TRL*/LRM*. (V ector

error correction reduces the e ects of system directivity , frequency response , source and

load match, and crosstalk.)

P ower meter calibration that allows you to use an HP-IB compatible power meter to

monitor and correct the analyzer's output power at each data point. (The analyzer stores a

power correction table that contains the correction values .)

Printing, Plotting, and Saving

Direct print or plot output of displayed measurement results , with a time stamp if desired,

to a compatible peripheral with a serial, parallel, or HP-IB interface .

Instrument states storage in internal memory for the following times , or on disk

inde nitely .

T emperature at 70

C : 250 days (0.68 year) characteristically

T emperature at 40

C : 1244 days (3.4 years) characteristically

T emperature at 25

C : 10 years characteristically

LIF/DOS disk formats for saving instrument states and measurement data.

Integration of a high capacity micro- oppy disk drive .

HP 8753D Description and Options 1-3

Show 4 more labels / short text on this page
A ccuracy · · ·
Page 39 of the Agilent Technologies 8753D user manual — Chapter 1 8753D Description and Options
Page 39 · click to enlarge
Page 39

Front P anel F eatures

Agilent Technologies 8753D — Chapter 1 8753D Description and Options diagram (page 39)
Fig. 6 · Chapter 1 8753D Description and Options

Figure 1-1. HP 8753D Front P anel

Figure 1-1 shows the location of the following front panel features and key function blocks .

These features are described in more detail later in this chapter , and in Chapter 9, \Key

De nitions ."

1. LINE switch. This switch controls ac power to the analyzer . 1 is on, 0 is o .
2. Display . This shows the measurement data traces , measurement annotation, and softkey

labels . The display is divided into speci c information areas , illustrated in Figure 1-2.

3. Softkeys . These keys provide access to menus that are shown on the display .
4. STIMULUS function block. The keys in this block allow you to control the analyzer

source's frequency , power , and other stimulus functions .

5. RESPONSE function block. The keys in this block allow you to control the measurement

and display functions of the active display channel.

6. A CTIVE CHANNEL keys . The analyzer has two independent display channels . These

keys allow you to select the active channel. Then any function you enter applies to this

active channel.

1-4 HP 8753D Description and Options

Page 40 of the Agilent Technologies 8753D user manual — . PORT 2
Page 40 · click to enlarge
Page 40
7. The ENTRY block. This block includes the knob , the step

keys , and the number

pad. These allow you to enter numerical data and control the markers .

Y ou can use the numeric keypad to select digits , decimal points , and a minus sign for

numerical entries . Y ou must also select a units terminator to complete value inputs .

8. INSTRUMENT ST A TE function block. These keys allow you to control

channel-independent system functions such as the following:

copying, save/recall, and HP-IB controller mode

frequency o set mode

harmonic measurements (Option 002)

time domain transform (Option 010)

HP-IB ST A TUS indicators are also included in this block.

9.

key . This key returns the instrument to either a known factory preset state , or

PRESET

a user preset state that can be de ned. Refer to Chapter 12, \Preset State and Memory

Allocation," for a complete listing of the instrument preset condition.

10. PORT 1 and PORT 2. These ports output a signal from the source and receive input

signals from a device under test. PORT 1 allows you to measure S

. PORT 2

allows you to measure S

11. PROBE PO WER connector . This connector (fused inside the instrument) supplies power

to an active probe for in-circuit measurements of ac circuits .

12. R CHANNEL connectors . These connectors allow you to apply an input signal to the

analyzer's R channel, for frequency o set mode .

13. Disk drive . This 3.5 inch drive allows you to store and recall instrument states and

measurement results for later analysis .

HP 8753D Description and Options 1-5

Show 8 more labels / short text on this page
* · + · limit testing · external source mode · tuned receiver mode · test sequence function · and S · and S
Page 41 of the Agilent Technologies 8753D user manual — F ORMA T
Page 41 · click to enlarge
Page 41
Agilent Technologies 8753D — Chapter 1 8753D Description and Options diagram (page 41)
Fig. 7 · Chapter 1 8753D Description and Options

Figure 1-2. Analyzer Display (Single Channel, Cartesian F ormat)

The analyzer display shows various measurement information:

The grid where the analyzer plots the measurement data.

The currently selected measurement parameters .

The measurement data traces .

Figure 1-2 illustrates the locations of the di erent information labels described below .

In addition to the full-screen display shown in Figure 1-2, a split display is available , as

described in Chapter 2, \Making Measurements ." In the split display mode , the analyzer

provides information labels for each half of the display .

Several display formats are available for di erent measurements , as described under

F ORMA T

" in Chapter 9, \Key De nitions ."

1. Stimulus Start V alue . This value could be any one of the following:

The start frequency of the source in frequency domain measurements .

The start time in CW mode (0 seconds) or time domain measurements .

The lower power value in power sweep .

When the stimulus is in center/span mode , the center stimulus value is shown in this

space .

1-6 HP 8753D Description and Options

Analyzer Display · \

Page 42 of the Agilent Technologies 8753D user manual — SCALE REF
Page 42 · click to enlarge
Page 42
2. Stimulus Stop V alue . This value could be any one of the following:

The stop frequency of the source in frequency domain measurements .

The stop time in time domain measurements or CW sweeps .

The upper limit of a power sweep .

When the stimulus is in center/span mode , the span is shown in this space . The stimulus

values can be blanked, as described under \

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FREQUENCY BLANK Key" in Chapter 9, \Key

De nitions ."

(F or CW time and power sweep measurements , the CW frequency is displayed centered

between the start and stop times or power values .)

3. Status Notations . This area shows the current status of various functions for the active

channel.

The following notations are used:

A vg = Sweep-to-sweep averaging is on. The averaging count is shown immediately

below . (See \

Key" in Chapter 9, \Key De nitions .")

A VG

Cor = Error correction is on. (F or error-correction procedures , refer to Chapter 5,

\Optimizing Measurement Results ." F or error correction theory , refer to Chapter

6, \Application and Operation Concepts .")

C? = Stimulus parameters have changed from the error-corrected state , or interpolated

error correction is on. (F or error-correction procedures , refer to Chapter 5,

\Optimizing Measurement Results ." F or error correction theory , refer to Chapter

6, \Application and Operation Concepts .")

C2 = Full two-port error-correction is active and either the power range for each port

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

is di erent (uncoupled), or the

TESTSET SW HOLD is activated. The annotation

occurs because the analyzer does not switch between the test ports every sweep

under these conditions . The measurement stays on the active port after an

initial cycling between the ports . (The active port is determined by the selected

measurement parameter .) Y ou can update all the parameters by pressing

MENU

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MEASURE RESTART ,or

MEAS

key .

Del = Electrical delay has been added or subtracted, or port extensions are active .

(See Chapter 6, \Application and Operation Concepts"and \

SCALE REF

Chapter 9, \Key De nitions .")

ext = W aiting for an external trigger .

Ofs = Frequency o set mode is on. (See \Frequency O set Operation" in Chapter 6,

\Application and Operation Concepts .")

Of? = Frequency o set mode error , the IF frequency is not within 10 MHz of expected

frequency . LO inaccuracy is the most likely cause . (See \Frequency O set

Operation" in Chapter 6, \Application and Operation Concepts .")

Gat = Gating is on (time domain Option 010 only). (F or time domain measurement

procedures , refer to Chapter 2, \Making Measurements ." F or time domain theory ,

refer to Chapter 6 \Application and Operation Concepts .")

HP 8753D Description and Options 1-7

· Key" in

Page 43 of the Agilent Technologies 8753D user manual — SCALE REF
Page 43 · click to enlarge
Page 43

H=2 = Harmonic mode is on, and the second harmonic is being measured (harmonics

Option 002 only). See \Analyzer Options A vailable" later in this chapter .)

H=3 = Harmonic mode is on, and the third harmonic is being measured (harmonics

Option 002 only). (See \Analyzer Options A vailable" later in this chapter .)

Hold sweep . (See

NNNNNNNNNNNNNN

HOLD in Chapter 9, \Key De nitions .")

man = W aiting for manual trigger .

PC = P ower meter calibration is on. (F or power meter calibration procedures , refer

to Chapter 5, \Optimizing Measurement Results ." F or power meter calibration

theory , refer to Chapter 6, \Application and Operation Concepts .")

PC? = The analyzer's source could not be set to the desired level, following a power

meter calibration. (F or power meter calibration procedures , refer to Chapter 5,

\Optimizing Measurement Results ." F or power meter calibration theory , refer to

Chapter 6, \Application and Operation Concepts .")

P? = Source power is unleveled at start or stop of sweep . (Refer to the

HP 8753D Network Analyzer Service Guide for troubleshooting.)

P# = Source power has been automatically set to minimum, due to receiver overload.

NNNNNNNNNNNNNNNNN

POWER in Chapter 9, \Key De nitions .")

PRm = P ower range is in manual mode .

Smo = Trace smoothing is on. (See \

" in Chapter 9, \Key De nitions ,")

A VG

tsH = Indicates that the test set hold mode is engaged.

That is , a mode of operation is selected which would cause repeated switching of

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

the step attenuator . This hold mode may be overridden. See

MEASURE RESTART

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NUMBER OF GROUPS in Chapter 9, \Key De nitions ."

" = F ast sweep indicator . This symbol is displayed in the status notation block when

sweep time is less than 1.0 second. When sweep time is greater than 1.0 second,

this symbol moves along the displayed trace .

*= Source parameters changedmeasured data in doubt until a complete fresh

sweep has been taken.

4. A ctive Entry Area. This displays the active function and its current value .
5. Message Area. This displays prompts or error messages .
6. Title . This is a descriptive alpha-numeric string title that you de ne and enter through

an attached keyboard or as described in Chapter 4, \Printing, Plotting, and Saving

Measurement Results ."

7. A ctive Channel. This is the number of the current active channel, selected with the

CHAN 1 and

CHAN 2 keys . If dual channel is on with an overlaid display , both channel 1

and channel 2 appear in this area.

8. Measured Input(s). This shows the S-parameter , input, or ratio of inputs currently

measured, as selected using the

MEAS

key . Also indicated in this area is the current

display memory status .

9. F ormat. This is the display format that you selected using the

F ORMA T

key .

10. Scale/Div . This is the scale that you selected using the

SCALE REF

key , in units

appropriate to the current measurement.

1-8 HP 8753D Description and Options

Show 4 more labels / short text on this page
Hld = · (See · or ·
Page 44 of the Agilent Technologies 8753D user manual — MARKER F CTN
Page 44 · click to enlarge
Page 44
11. Reference Level. This value is the reference line in Cartesian formats or the outer circle

in polar formats , whichever you selected using the

SCALE REF

key . The reference level is

also indicated by a small triangle adjacent to the graticule , at the left for channel 1 and at

the right for channel 2 in Cartesian formats .

12. Marker V alues . These are the values of the active marker , in units appropriate to the

current measurement. (Refer to \Using Analyzer Display Markers" in Chapter 2, \Making

Measurements .")

13. Marker Stats , Bandwidth. These are statistical marker values that the analyzer

calculates when you access the menus with the

MARKER F CTN

key . (Refer to \Using

Analyzer Display Markers" in Chapter 2, \Making Measurements .")

14. Softkey Labels . These menu labels rede ne the function of the softkeys that are located

to the right of the analyzer display .

15. P ass F ail. During limit testing, the result will be annunciated as PASS if the limits are not

exceeded, and FAIL if any points exceed the limits .

HP 8753D Description and Options 1-9

Page 45 of the Agilent Technologies 8753D user manual — Chapter 1 8753D Description and Options
Page 45 · click to enlarge
Page 45

Rear P anel F eatures and Connectors

Agilent Technologies 8753D — Chapter 1 8753D Description and Options diagram (page 45)
Fig. 8 · Chapter 1 8753D Description and Options

Figure 1-3. HP 8753D Rear P anel

Figure 1-3 illustrates the features and connectors of the rear panel, described below .

Requirements for input signals to the rear panel connectors are provided in Chapter 7,

\Speci cations and Measurement Uncertainties ."

1. Serial number plate . The serial number of the instrument is located on this plate .
2. EXTERNAL MONITOR: RED , GREEN, BLUE Video output connectors provide analog

red, green, and blue video signals which you can use to drive an analog multi-sync

external monitor . The monitor must be compatible with the analyzer's 25.5 kHz scan rate

and video levels: 1 Vp-p , 0.7 V=white , 0 V=black, 0.3 V sync , sync on green.

3. HP-IB connector . This allows you to connect the analyzer to an external controller ,

compatible peripherals , and other instruments for an automated system. Refer to Chapter

11, \Compatible P eripherals ," in this document for HP-IB information, limitations , and

con gurations .

4. P ARALLEL interface . This connector allows the analyzer to output to a peripheral

with a parallel input. Also included, is a general purpose input/output (GPIO) bus that

can control eight output bits and read ve input bits through test sequencing. Refer to

Chapter 11, \Compatible P eripherals ," for information on con guring a peripheral. Also

refer to \Application and Operation Concepts" for information on GPIO .

5. RS-232 interface . This connector allows the analyzer to output to a peripheral with an

RS-232 (serial) input.

6. KEYBO ARD input (DIN). This connector allows you to connect an external keyboard.

This provides a more convenient means to enter a title for storage les , as well as

substitute for the analyzer's front panel keyboard.

7. P ower cord receptacle , with fuse . F or information on replacing the fuse , refer to the

HP 8753D Network Analyzer Installation and Quick Start Guide or the

HP 8753D Network Analyzer Service Guide.

8. Line voltage selector switch. F or more information refer to the HP 8753D Network

Analyzer Installation and Quick Start Guide.

9. 10 MHZ REFERENCE ADJUST . (Option 1D5)

1-10 HP 8753D Description and Options

Page 46 of the Agilent Technologies 8753D user manual — Chapter 1 8753D Description and Options
Page 46 · click to enlarge
Page 46
10. 10 MHZ PRECISION REFERENCE OUTPUT . (Option 1D5)
11. EXTERNAL REFERENCE INPUT connector . This allows for a frequency reference signal

input that can phase lock the analyzer to an external frequency standard for increased

frequency accuracy .

The analyzer automatically enables the external frequency reference feature when a

signal is connected to this input. When the signal is removed, the analyzer automatically

switches back to its internal frequency reference .

12. A UXILIARY INPUT connector . This allows for a dc or ac voltage input from an external

signal source , such as a detector or function generator , which you can then measure , using

the S-parameter menu. (Y ou can also use this connector as an analog output in service

routines , as described in the service manual.)

13. EXTERNAL AM connector . This allows for an external analog signal input that is

applied to the ALC circuitry of the analyzer's source . This input analog signal amplitude

modulates the RF output signal.

14. EXTERNAL TRIGGER connector . This allows connection of an external negative-going

TTL-compatible signal that will trigger a measurement sweep . The trigger can be set to

external through softkey functions .

15. TEST SEQUENCE. This outputs a TTL signal that can be programmed in a test sequence

to be high or low , or pulse (10 seconds) high or low at the end of a sweep for robotic part

handler interface .

16. LIMIT TEST . This outputs a TTL signal of the limit test results as follows:
P assTTL high
F ailTTL low
17. BIAS INPUTS AND FUSES . These connectors bias devices connected to port 1 and

port 2. The fuses (1 A, 125 V) protect the port 1 and port 2 bias lines .

18. TEST SET INTERCONNECT . This allows you to connect an HP 8753D Option 011

analyzer to an HP 85046A/B or 85047A S-parameter test set using the interconnect

cable supplied with the test set. The S-parameter test set is then fully controlled by the

analyzer .

19. F an. This fan provides forced-air cooling for the analyzer .

HP 8753D Description and Options 1-11

Page 47 of the Agilent Technologies 8753D user manual — Chapter 1 8753D Description and Options
Page 47 · click to enlarge
Page 47

Analyzer Options A vailable

Option 1D5, High Stability Frequency Reference

Option 1D5 o ers 0.05 ppm temperature stability from 0 to 60

C (referenced to 25

C).

Option 002, Harmonic Mode

Provides measurement of second or third harmonics of the test device's fundamental output

signal. Frequency and power sweep are supported in this mode . Harmonic frequencies can be

measured up to the maximum frequency of the receiver . However , the fundamental frequency

may not be lower than 16 MHz.

Option 006, 6 GHz Operation

Option 006 extends the maximum source and receiver frequency of the analyzer to 6 GHz.

Option 010, Time Domain

This option displays the time domain response of a network by computing the inverse F ourier

transform of the frequency domain response . It shows the response of a test device as a

function of time or distance . Displaying the re ection coecient of a network versus time

determines the magnitude and location of each discontinuity . Displaying the transmission

coecient of a network versus time determines the characteristics of individual transmission

paths . Time domain operation retains all accuracy inherent with the correction that is active in

the frequency domain. The time domain capability is useful for the design and characterization

of such devices as SA W lters , SA W delay lines , RF cables , and RF antennas .

Option 011, Receiver Con guration

Option 011 allows front panel access to the R, A, and B samplers and receivers . The transfer

switch, couplers , and bias tees have been removed. Therefore , external accessories are

required to make most measurements .

Option 075, 75 Impedance

Option 075 o ers 75 ohm impedance bridges with type-N test port connectors .

Option 1CM, Rack Mount Flange Kit Without Handles

Option 1CM is a rack mount kit containing a pair of anges and the necessary hardware to

mount the instrument, with handles detached, in an equipment rack with 482.6 mm (19 inches)

horizontal spacing.

Option 1CP , Rack Mount Flange Kit With Handles

Option 1CP is a rack mount kit containing a pair of anges and the necessary hardware to

mount the instrument with handles attached in an equipment rack with 482.6 mm (19 inches)

spacing.

1-12 HP 8753D Description and Options

·

Page 48 of the Agilent Technologies 8753D user manual — Chapter 1 8753D Description and Options
Page 48 · click to enlarge
Page 48

Service and Support Options

The analyzer automatically includes a one-year on-site service warranty , where available . The

following service and support products are available with an HP 8753D network analyzer at

any time during or after the time of purchase . A dditional service and support options may be

available at some sites . Consult your local HP customer engineer for details .

On-Site System V eri cation (+23G)

On-site system veri cation (performed by a Hewlett-P ackard customer engineer), con rms

the system's error-corrected uncertainty performance by measuring traceable 7 mm devices .

It provides a hardcopy listing of both ideal and actual data, together with a certi cate of

traceability . Preventive maintenance is performed at the time of system veri cation. Travel

through Zone 3 (up to 100 miles/160 km from Hewlett-P ackard's nearest service-responsible

oce) is included.

Standard System Maintenance Service (+02A)

This option provides four-hour , on-site response through Travel Zone 3 on all service requests

for the HP 8753D (and a 50 ohm test set for Option 011), by a Hewlett-P ackard customer

engineer .

Basic System Maintenance Service (+02B)

This option provides next day on-site response through Travel Zone 3 on all service requests for

the HP 8753D (and a 50 ohm test set for Option 011), by a Hewlett-P ackard customer engineer .

Return to HP Full Service Agreement (+22A)

This option is a one-year service contract for any repair of the HP 8753D at a Hewlett-P ackard

repair facility . One complete calibration procedure is included.

Return to HP Repair Agreement (+22B)

This option provides repair of the HP 8753D at a Hewlett-P ackard repair facility for one year .

F ollowing repair , the instrument is tested functionally but is not fully calibrated.

Return to HP Calibration Agreement (+22C)

This option provides a once-a-year complete calibration procedure at a Hewlett-P ackard facility .

Return to HP Calibration (+22G)

This option is a one-time complete calibration procedure performed at a Hewlett-P ackard

facility . The procedure veri es that the HP 8753D is performing according to its published

speci cations .

HP 8753D Description and Options 1-13

Page 49 of the Agilent Technologies 8753D user manual — Chapter 1 8753D Description and Options
Page 49 · click to enlarge
Page 49

Changes between the HP 8753 Network Analyzers

T able 1-1. Comparing the HP 8753A/B/C/D

F eature 8753A 8753B 8753C 8753D 8753D Opt

Fully integrated measurement system (built-in

No No No Y es No

T est port power range (dBm) y y y +10 to 85 y

A uto/manual power range selecting No No No Y es No

P ort power coupling/uncoupling No No No Y es No

Internal disk drive No No No Y es Y es

Precision frequency reference (Option 1D5) No No No Y es Y es

Frequency range - low end 300 kHz 300 kHz 300 kHz 30 kHz 30/300 kHz*

Ext. freq. range to 6 GHz (Option 006) No Y es Y es Y es Y es

75 system impedance (Option 075) y y y Y es y

TRL*/LRM* correction No No No Y es Y es

P ower meter calibration No Y es Y es Y es Y es

Interpolated error correction No Y es Y es Y es Y es

Max. error corrected measurement points 801 1601

Segmented error correction in freq. list mode No No Y es Y es Y es

Color CRT No No Y es Y es Y es

T est sequencing No Y es Y es Y es Y es

A utomatic sweep time No Y es Y es Y es Y es

External source capability No Y es Y es Y es Y es

Tuned receiver mode No Y es Y es Y es Y es

Printer/plotter bu er No Y es Y es Y es Y es

Harmonic measurements (Option 002) No Y es Y es Y es Y es

Frequency o set mode (mixer measurements) No Y es Y es Y es Y es

dc bias to test device y y y Y es y

InterfacesRS-232, parallel, and DIN

No No No Y es Y es

User-de ned preset No No No Y es Y es

Non-volatile memory 16 Kbytes 16 Kbytes 16 Kbytes 512 Kbytes 512 Kbytes

30 kHz to 3 GHz 100 dB 100 dB 100 dB 110 dBz 100 dB

3 GHz to 6 GHz N/A 80 dB 80 dB 105 dB 110 dB

Real time clock No No No Y es Y es

* 300 kHz to 3 GHz, without Option 006, or 30 kHz to 6 GHz, with Option 006.

y F or this network analyzer , the feature is dependent on the test set being used.

z 90 dB from 30 kHz to 50 kHz, 100 dB from 300 kHz to 16 MHz.

1-14 HP 8753D Description and Options

test set) · keyboard · Dynamic range

Page 50 of the Agilent Technologies 8753D user manual — Chapter 2 Making Measurements
Page 50 · click to enlarge
Page 50

This Chapter contains the following example procedures for making measurements or using

particular functions:

Basic measurement sequence and example

Magnitude and insertion phase response

Electrical length and phase distortion

Deviation from linear phase

Gain and reverse isolation

High P ower Measurements

Where to Look for More Information

A dditional information about many of the topics discussed in this Chapter is located in the

following areas:

Chapter 4, \Printing, Plotting, and Saving Measurement Results ," contains instructions

for saving to disk or the analyzer internal memory , and printing and plotting displayed

measurements .

Chapter 5, \Optimizing Measurement Results ," describes techniques and functions for

achieving the best measurement results .

Chapter 6, \Application and Operation Concepts ," contains explanatory-style information

about many applications and analyzer operation.

Chapter 9, \Key De nitions ," describes all the front panel keys and softkeys .

Chapter 11, \Compatible P eripherals ," lists measurement and system accessories , and other

applicable equipment compatible with the analyzer .

Show 15 more labels / short text on this page
Making Measurements · Setting frequency range · Setting source power · Analyzer display functions · Analyzer marker functions · Group delay · Limit testing · Gain compression · Tuned Receiver Mode · T est sequencing · Time domain · Transmission response · Re ection response · Non-Coaxial Measurements · Making Measurements 2-1
Page 51 of the Agilent Technologies 8753D user manual — Chapter 2 Making Measurements
Page 51 · click to enlarge
Page 51

Principles of Microwave Connector Care

Proper connector care and connection techniques are critical for accurate , repeatable

measurements .

Refer to the calibration kit documentation for connector care information. Prior to making

connections to the network analyzer , carefully review the information about inspecting,

cleaning and gaging connectors .

Having good connector care and connection techniques extends the life of these devices . In

addition, you obtain the most accurate measurements .

This type of information is typically located in Chapter 3 of the calibration kit manuals .

F or additional connector care instruction, contact your local Hewlett-P ackard Sales and Service

Oce about course numbers HP 85050A+24A and HP 85050A+24D .

See the following table for quick reference tips about connector care .

T able 2-1. Connector Care Quick Reference

Keep connectors clean T ouch mating-plane surfaces

Extend sleeve or connector nut Set connectors contact-end down

Use plastic end-caps during storage

Inspect all connectors carefully Use a damaged connector - ever

Look for metal particles , scratches , and dents

Try compressed air rst Use any abrasives

Use isopropyl alcohol Get liquid into plastic support beads

Clean and zero the gage before use Use an out-of-spec connector

Use the correct gage type

Use correct end of calibration block

Gage all connectors before rst use

Align connectors carefully Apply bending force to connection

Make preliminary connection lightly Over tighten preliminary connection

Turn only the connector nut Twist or screw any connection

Use a torque wrench for nal connect Tighten past torque wrench \break" point

Show 12 more labels / short text on this page
Handling and Storage · Do Do Not · Visual Inspection · Do Do Not · Connector Cleaning · Do Do Not · Clean connector threads · Gaging Connectors · Do Do Not · Making Connections · Do Do Not · 2-2 Making Measurements
Page 52 of the Agilent Technologies 8753D user manual — CENTER
Page 52 · click to enlarge
Page 52

Basic Measurement Sequence and Example

There are ve basic steps when you are making a measurement.

1. Connect the device under test and any required test equipment.

Damage may result to the device under test if it is sensitive to analyzer's

default output power level. T o avoid damaging a sensitive DUT , perform step 2

before step 1.

2. Choose the measurement parameters .
3. P erform and apply the appropriate error-correction.
4. Measure the device under test.
5. Output the measurement results .

This example procedure shows you how to measure the transmission response of a bandpass

lter .

1. Connect the device under test and any required test equipment.
1. Make the connections as shown in Figure 2-1.
Agilent Technologies 8753D — Chapter 2 Making Measurements diagram (page 52)
Fig. 9 · Chapter 2 Making Measurements

Figure 2-1. Basic Measurement Setup

2. Choose the measurement parameters .
2. Press

PRESET

.

T o set preset to \F actory Preset," press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRESET: FACTORY

PRESET

Setting the Frequency Range .

3. T o set the center frequency to 134 MHz, press:

CENTER

M/

Show 8 more labels / short text on this page
Basic Measurement Sequence · Basic Measurement Example · · · · 134 · · Making Measurements 2-3
Page 53 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 53 · click to enlarge
Page 53
4. T o set the span to 30 MHz, press:

SP AN

M/

Y ou could also press the

ST ART

STOP

keys and enter the frequency range

limits as start frequency and stop frequency values .

Setting the Source P ower .

5. T o change the power level to 5 dBm, press:

NNNNNNNNNNNNNNNNN

POWER

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Y ou could also press

POWER RANGE MAN

POWER RANGES and select one of the

power ranges to keep the power setting within the de ned range .

Setting the Measurement.

6. T o change the number of measurement data points to 101, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NUMBER OF POINTS

MENU

7. T o select the transmission measurement, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Trans:FWD S21 (B/R)

MEAS

8. T o view the data trace , press:

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNN

SCALE REF

AUTOSCALE

3. P erform and apply the appropriate error-correction.
9. Refer to the \Optimizing Measurement Results" Chapter for procedures on correcting

measurement errors .

10. T o save the instrument state and error-correction in the analyzer internal memory , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SELECT DISK

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INTERNAL MEMORY

NNNNNNNNNNNNNNNNNNNN

RETURN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SAVE STATE

SA VE/RECALL

4. Measure the device under test.
11. Replace any standard used for error-correction with the device under test.
12. T o measure the insertion loss of the bandpass lter , press:

MARKER

M/

5. Output the measurement results .
13. T o create a hardcopy of the measurement results , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNN

COPY

PLOT )

Refer to Chapter 4, \Printing, Plotting, and Saving Measurement Results ," for procedures

on how to de ne a print, plot, or save . F or information on con guring a peripheral, refer

to Chapter 11, \Compatible P eripherals ."

Show 27 more labels / short text on this page
· · 30 · · and · · · · · · · · + · · · · · · · · · 134 · · · · PRINT MONOCHROME (or · 2-4 Making Measurements
Page 54 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 54 · click to enlarge
Page 54

Using the Display Functions

T o View Both Measurement Channels

In some cases , you may want to view more than one measured parameter at a time .

Simultaneous gain and phase measurements for example , are useful in evaluating stability in

negative feedback ampli ers . Y ou can easily make such measurements using the dual channel

display .

1. T o see both channels simultaneously , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DISPLA Y

DUAL CHAN ON

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 54)
Fig. 10

Figure 2-2. Example of Viewing Both Channels Simultaneously

· · Making Measurements 2-5

Page 55 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 55 · click to enlarge
Page 55

Y ou can view the measurements on separate displays , press:

NNNNNNNNNNNNNN

MORE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SPLIT DISP ON The

analyzer shows channel 1 on the upper half of the display and channel 2 on the lower half

of the display . The analyzer also defaults to measuring S

on channel 1 and S

Agilent Technologies 8753D — NNNNNNNNNNNNNN diagram (page 55)
Fig. 11

Figure 2-3. Example Dual Channel With Split Display On

T o return to one display , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SPLIT DISPLAY OFF

Y ou can control the stimulus functions of the two channels independent of each

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

other , by pressing

MENU

COUPLED CH OFF .

T o Save a Data Trace to the Display Memory

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DISPLA Y

DATA ! MEMORY to store the current active measurement data in the memory of

the active channel.

The data trace is now also the memory trace . Y ou can use a memory trace for subsequent math

manipulations .

T o View the Measurement Data and Memory Trace

The analyzer default setting shows you the current measurement data for the active channel.

1. T o view a data trace that you have already stored to the active channel memory , press:

NNNNNNNNNNNNNNNNNNNN

MEMORY

DISPLA Y

This is the only memory display mode where you can change the smoothing and gating of

the memory trace .

2. T o view both the memory trace and the current measurement data trace , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DISPLA Y

Show 10 more labels / short text on this page
on channel · · Press · · · · · · DATA and MEMORY · 2-6 Making Measurements
Page 56 of the Agilent Technologies 8753D user manual — DISPLA Y
Page 56 · click to enlarge
Page 56

T o Divide Measurement Data by the Memory Trace

Y ou can use this feature for ratio comparison of two traces , for example , measurements of gain

or attenuation.

1. Y ou must have already stored a data trace to the active channel memory , as described in \T o

Save a Data Trace to the Display Memory ."

N

NNNNNNNNNNNNNNNNNNNNNNNNN

DISPLA Y

DATA/MEM to divide the data by the memory .

The analyzer normalizes the data to the memory , and shows the results .

T o Subtract the Memory Trace from the Measurement Data Trace

Y ou can use this feature for storing a measured vector error , for example , directivity . Then,

you can later subtract it from the device measurement.

1. Y ou must have already stored a data trace to the active channel memory , as described in \T o

Save a Data Trace to the Display Memory ."

NNNNNNNNNNNNNNNNNNNNNNNNNN

DATA-MEM to subtract the memory from the measurement data.

DISPLA Y

The analyzer performs a vector subtraction on the complex data.

T o Ratio Measurements in Channel 1 and 2

Y ou may want to use this feature when making ampli er measurements to produce a trace that

represents gain compression. F or example , with the channels uncoupled, you can increase the

power for channel 2 while channel 1 remains unchanged. This will allow you to observe the

gain compression on channel 2 .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MENU

COUPLED CH OFF to uncouple the channels .

2. Make sure that both channels must have the same number of points .

a.

CHAN 1

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MENU

NUMBER OF POINTS and notice the number of points setting, shown

on the analyzer display .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

b.

CHAN 2

MENU

NUMBER OF POINTS and enter the same value that you observed

for the channel 1 setting.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DUAL CHAN ON

NNNNNNNNNNNNNN

MORE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

D2/D1 TO D2 ON to ratio channels 1 and 2, and put

DISPLA Y

the results in the channel 2 data array . This ratio is applied to the complex data.

4. Refer to Chapter \Measuring Gain Compression" for the procedure on identifying the 1 dB

compression point.

Show 13 more labels / short text on this page
Press · · Press · · Press · · Press · · Press · · Press · · Making Measurements 2-7
Page 57 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNN
Page 57 · click to enlarge
Page 57

T o Title the A ctive Channel Display

NNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNN

TITLE to access the title menu.

DISPLA Y

MORE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ERASE TITLE and enter the title you want for your measurement display .

If you have a DIN keyboard attached to the analyzer , type the title you want from the

keyboard. Then press

ENTER

to enter the title into the analyzer . Y ou can enter a title

that has a maximum of 50 characters .

If you do not have a DIN keyboard attached to the analyzer , enter the title from the

analyzer front panel.

a. Turn the front panel knob to move the arrow pointer to the rst character of the title .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

b.

SELECT LETTER .

c. Repeat the previous two steps to enter the rest of the characters in your title . Y ou can

enter a title that has a maximum of 50 characters .

d. Press

NNNNNNNNNNNNNN

DONE to complete the title entry .

Agilent Technologies 8753D — NNNNNNNNNNNNNN diagram (page 57)
Fig. 12

Figure 2-4. Example of a Display Title

Show 5 more labels / short text on this page
Press · · Press · Press · 2-8 Making Measurements
Page 58 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 58 · click to enlarge
Page 58

Using Analyzer Display Markers

The analyzer markers provide numerical readout of trace data. Y ou can control the marker

search, the statistical functions , and the capability for quickly changing stimulus parameters

with markers , from the

key .

MARKER F CTN

Markers have a stimulus value (the x-axis value in a Cartesian format) and a response value

(the y-axis value in a Cartesian format). In a polar or Smith chart format, the second part of

a complex data pair is also provided as an auxiliary response value . When you switch on a

marker , and no other function is active , the analyzer shows the marker stimulus value in the

active entry area. Y ou can control the marker with the front panel knob , the step keys , or the

front panel numeric keypad.

If you activate both data and memory traces , the marker values apply to the data trace .

If you activate only the memory trace , the marker values apply to the memory trace .

If you activate a memory math function (data/memory or data-memory), the marker values

apply to the trace resulting from the memory math function.

The examples in this section are shown with lter measurement results . The measurement

parameters are set as follows:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TransFWD S21 (B/R)

MEAS

CENTER

M/

SP AN

M/

T o Use Continuous and Discrete Markers

The analyzer can either place markers on discrete measured points , or move the markers

continuously along a trace by interpolating the data value between measured points .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER MODE MENU and select one of the following choices:

MARKER F CTN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKERSCONTINUOUS if you want the analyzer to place markers at any point

on the trace , by interpolating between measured points . This default mode allows you to

conveniently obtain round numbers for the stimulus value .

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKERSDISCRETE if you want the analyzer to place markers only on measured

trace points determined by the stimulus settings . This may be the best mode to use with

automated testing, using a computer or test sequencing because the analyzer does not

interpolate between measured points .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKERSDISCRETE will also a ect marker search and positioning

functions when the value entered in a search or positioning function does not

exist as a measurement point.

Show 16 more labels / short text on this page
· · · · 134 · · · · 25 · · Press · · Choose · Choose · Using · Making Measurements 2-9
Page 59 of the Agilent Technologies 8753D user manual — MARKER 5
Page 59 · click to enlarge
Page 59

T o A ctivate Display Markers

T o switch on marker 1 and make it the active marker , press:

N

MARKER 1

NNNNNNNNNNNNNNNNNNNNNNNNN

MARKER

The active marker appears on the analyzer display as r . The active marker stimulus value is

displayed in the active entry area. Y ou can modify the stimulus value of the active marker ,

using the front panel knob or numerical keypad. All of the marker response and stimulus

values are displayed in the upper right corner of the display .

Agilent Technologies 8753D — N diagram (page 59)
Fig. 13

Figure 2-5. A ctive Marker Control Example

T o switch on the corresponding marker and make it the active marker , press:

NNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER 2 ,

MARKER 3 ,

MARKER 4 , or

MARKER 5

All of the markers , other than the active marker , become inactive and are represented on the

analyzer display as .

Agilent Technologies 8753D — MARKER 5 diagram (page 59)
Fig. 14

Figure 2-6. A ctive and Inactive Markers Example

· · 2-10 Making Measurements

Page 60 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 60 · click to enlarge
Page 60

T o switch o all of the markers , press:

NNNNNNNNNNNNNNNNNNNNNNN

ALL OFF

T o Use Delta () Markers

This is a relative mode , where the marker values show the position of the active marker

relative to the delta reference marker . Y ou can switch on the delta mode by de ning one of the

ve markers as the delta reference .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNN

1. Press

MARKER

MODE MENU

REF=1 to make marker 1 a reference marker .

2. T o move marker 1 to any point that you want to reference:

a. Turn the front panel knob .

OR

a. Enter the frequency value (relative to the reference marker) on the numeric keypad.

NNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER 2 and move marker 2 to any position that you want to measure in reference

to marker 1.

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 60)
Fig. 15

Figure 2-7. Marker 1 as the Reference Marker Example

4. T o change the reference marker to marker 2, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNN

MODE MENU

REF=2

T o A ctivate a Fixed Marker

When a reference marker is xed, it does not rely on a current trace to maintain its xed

position. The analyzer allows you to activate a xed marker with one of the following key

sequences:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER

MODE MENU

REF=FIXED MKR

NNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER

MKR ZERO

Show 7 more labels / short text on this page
· Press · · · · · Making Measurements 2-11
Page 61 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 61 · click to enlarge
Page 61

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REF=FIXED MKR K ey to activate a Fixed Reference Marker

1. T o set the frequency value of a xed marker that appears on the analyzer display , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER

MODE MENU

REF=FIXED MKR

MODE MENU

FIXED MKR POSITION

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FIXED MKR STIMULUS and turn the front panel knob or enter a value from the front panel

keypad.

The marker is shown on the display as a small delta (), smaller than the inactive marker

triangles .

2. T o set the response value (dB) of a xed marker , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FIXED MKR VALUE and turn the front panel knob or enter a value from the front panel

keypad.

In a Cartesian format the setting is the y-axis value . In polar or Smith chart format, with a

magnitude/phase marker , a real/imaginary marker , an R+jX marker , or a G+jB marker , the

setting applies to the rst part of the complex data pair . (Fixed marker response values are

always uncoupled in the two channels .)

3. T o set the auxiliary response value of a xed marker when you are viewing a polar or Smith

format, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FIXED MKR AUX VALUE and turn the front panel knob or enter a value from the front panel

keypad.

This value is the second part of complex data pair , and applies to a magnitude/phase marker ,

a real/imaginary marker , an R+jX marker , or a G+jB marker . (Fixed marker auxiliary

response values are always uncoupled in the two channels .)

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 61)
Fig. 16

Figure 2-8.

Example of a Fixed Reference Marker Using

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REF=FIXED MKR

Show 4 more labels / short text on this page
Using the · · · 2-12 Making Measurements
Page 62 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNN
Page 62 · click to enlarge
Page 62

NNNNNNNNNNNNNNNNNNNNNNNNNNNN

MKR ZERO K ey to A ctivate a Fixed Reference Marker

Marker zero enters the position of the active marker as the reference position. Alternatively ,

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

you can specify the xed point with

FIXED MKR POSITION . Marker zero is canceled by

switching delta mode o .

1. T o place marker 1 at a point that you would like to reference , press:

MARKER

and turn the front panel knob or enter a value from the front panel keypad.

2. T o measure values along the measurement data trace , relative to the reference point that

you set in the previous step , press:

NNNNNNNNNNNNNNNNNNNNNNNNNN

MKR ZERO and turn the front panel knob or enter a value from the front panel keypad.

3. T o move the reference position, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MODE MENU

FIXED MKR POSITION

FIXED MKR STIMULUS and turn the front panel knob

or enter a value from the front panel keypad.

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 62)
Fig. 17

Figure 2-9. Example of a Fixed Reference Marker Using

NNNNNNNNNNNNNNNNNNNNNNNNNN

MKR ZERO

Using the · · Making Measurements 2-13

Page 63 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNN
Page 63 · click to enlarge
Page 63

T o Couple and Uncouple Display Markers

At a preset state , the markers have the same stimulus values on each channel, but they can be

uncoupled so that each channel has independent markers .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER F CTN

MARKER MODE MENU and select from the following keys:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKERSCOUPLED if you want the analyzer to couple the marker stimulus values

for the two display channels .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKERSUNCOUPLED if you want the analyzer to uncouple the marker stimulus

values for the two display channels . This allows you to control the marker stimulus values

independently for each channel.

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 63)
Fig. 18

Figure 2-10. Example of Coupled and Uncoupled Markers

T o Use P olar F ormat Markers

The analyzer can display the marker value as magnitude and phase , or as a real/imaginary pair:

NNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNN

LIN MKR gives linear magnitude and phase ,

LOG MKR gives log magnitude and phase ,

gives the real value rst, then the imaginary value .

Y ou can use these markers only when you are viewing a polar display format. (The format is

available from the

F ORMA T

key .)

F or greater accuracy when using markers in the polar format, it is

recommended to activate the discrete marker mode . Press

MARKER F CTN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MKR MODE MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKERS:DISCRETE .

1. T o access the polar markers , press:

NNNNNNNNNNNNNNNNN

F ORMA T

POLAR

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER F CTN

MARKER MODE MENU

POLAR MKR MENU

Show 11 more labels / short text on this page
Press · · Choose · Choose · Re/Im · · · · · · 2-14 Making Measurements
Page 64 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 64 · click to enlarge
Page 64
2. Select the type of polar marker you want from the following choices:

NNNNNNNNNNNNNNNNNNNNNNN

LIN MKR if you want to view the magnitude and the phase of the active marker .

The magnitude values appear in units and the phase values appear in degrees .

NNNNNNNNNNNNNNNNNNNNNNN

LOG MKR if you want to view the logarithmic magnitude and the phase of the

active marker . The magnitude values appear in dB and the phase values appear in

degrees .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Re/Im MKR if you want to view the real and imaginary pair , where the complex

data is separated into its real part and imaginary part. The analyzer shows the rst

marker value the real part (M cos ), and the second value is the imaginary part

(M sin , where M=magnitude).

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 64)
Fig. 19

Figure 2-11. Example of a Log Marker in P olar F ormat

T o Use Smith Chart Markers

The amount of power re ected from a device is directly related to the impedance of the device

and the measuring system. Each value of the re ection coecient () uniquely de nes a device

impedance; = 0 only occurs when the device and analyzer impedance are exactly the same .

The re ection coecient for a short circuit is: = 1

. Every other value for also

corresponds uniquely to a complex device impedance , according to the equation:

Z

= [( 1 + ) / (1 )]Z

is your test device impedance and Z

is the measuring system's characteristic

impedance .

F or greater accuracy when using markers in the Smith chart format, it is

recommended to activate the discrete marker mode . Press

MARKER F CTN

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MKR MODE MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKERS:DISCRETE .

1. Press

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

F ORMA T

SMITH CHART .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER F CTN

MARKER MODE MENU

SMITH MKR MENU and turn the front panel knob or

enter a value from the front panel keypad to read the resistive and reactive components of

the complex impedance at any point along the trace . This is the default Smith chart marker .

Show 10 more labels / short text on this page
Choose · Choose · Choose · · where Z · · · Press · · Making Measurements 2-15
Page 65 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNN
Page 65 · click to enlarge
Page 65

The marker annotation tells that the complex impedance is capacitive in the bottom half of

the Smith chart display and is inductive in the top half of the display .

NNNNNNNNNNNNNNNNNNNNNNN

LIN MKR if you want the analyzer to show the linear magnitude and the phase of

the re ection coecient at the marker .

NNNNNNNNNNNNNNNNNNNNNNN

LOG MKR if you want the analyzer to show the logarithmic magnitude and the

phase of the re ection coecient at the active marker . This is useful as a fast method of

obtaining a reading of the log magnitude value without changing to log magnitude format.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Re/Im MKR if you want the analyzer to show the values of the re ection

coecient at the marker as a real and imaginary pair .

NNNNNNNNNNNNNNNNNNNNNNNNNN

R+

X MKR to show the real and imaginary parts of the device impedance at

the marker . Also shown is the equivalent series inductance or capacitance (the series

resistance and reactance , in ohms).

NNNNNNNNNNNNNNNNNNNNNNNNNN

G+

B MKR to show the complex admittance values of the active marker in

rectangular form. The active marker values are displayed in terms of conductance (in

Siemens), susceptance , and equivalent parallel circuit capacitance or inductance . Siemens

are the international unit of admittance and are equivalent to mhos (the inverse of ohms).

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 65)
Fig. 20

Figure 2-12. Example of Impedance Smith Chart Markers

T o Set Measurement P arameters Using Markers

The analyzer allows you to set measurement parameters with the markers , without going

through the usual key sequence . Y ou can change certain stimulus and response parameters to

make them equal to the current active marker value .

Show 8 more labels / short text on this page
Choose · Choose · Choose · Choose · j · Choose · j · 2-16 Making Measurements
Page 66 of the Agilent Technologies 8753D user manual — N
Page 66 · click to enlarge
Page 66

Setting the Start Frequency

1. Press

MARKER F CTN

and turn the front panel knob or enter a value from the front panel

keypad to position the marker at the value that you want for the start frequency .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER !START to change the start frequency value to the value of the active marker .

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 66)
Fig. 21

Figure 2-13. Example of Setting the Start Frequency Using a Marker

Setting the Stop Frequency

1. Press

and turn the front panel knob or enter a value from the front panel

MARKER F CTN

keypad to position the marker at the value that you want for the stop frequency .

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER !STOP to change the stop frequency value to the value of the active marker .

Agilent Technologies 8753D — N diagram (page 66)
Fig. 22

Figure 2-14. Example of Setting the Stop Frequency Using a Marker

Press · Press · Making Measurements 2-17

Page 67 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 67 · click to enlarge
Page 67

Setting the Center Frequency

1. Press

MARKER F CTN

and turn the front panel knob or enter a value from the front panel

keypad to position the marker at the value that you want for the center frequency .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER ! CENTER to change the center frequency value to the value of the active

marker .

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 67)
Fig. 23

Figure 2-15. Example of Setting the Center Frequency Using a Marker

Press · 2-18 Making Measurements

Page 68 of the Agilent Technologies 8753D user manual — MARKER F CTN
Page 68 · click to enlarge
Page 68

Setting the Frequency Span

Y ou can set the span equal to the spacing between two markers . If you set the center

frequency before you set the frequency span, you will have a better view of the area of

interest.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNN

1. Press

MARKER

MODE MENU

REF=1

MARKER 2 .

2. Turn the front panel knob or enter a value from the front panel keypad to position the

markers where you want the frequency span.

Iterate between marker 1 and marker 2 by pressing

NNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER 2 , respectively ,

and turning the front panel knob or entering values from the front panel keypad to position

the markers around the center frequency . When nished positioning the markers , make sure

that marker 2 is selected as the active marker .

NNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNN

Step 2 can also be performed using

MARKER 1 . However , when

using this method, it will not be possible to iterate between marker zero and

marker 1.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

3. Press

MARKER F CTN

MARKER ! SPAN to change the frequency span to the range between

marker 1 and marker 2.

Agilent Technologies 8753D — MARKER F CTN diagram (page 68)
Fig. 24

Figure 2-16. Example of Setting the Frequency Span Using Markerj

Show 5 more labels / short text on this page
· Marker 1 and · MKR ZERO and · · Making Measurements 2-19
Page 69 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 69 · click to enlarge
Page 69

Setting the Display Reference V alue

1. Press

MARKER F CTN

and turn the front panel knob or enter a value from the front panel

keypad to position the marker at the value that you want for the analyzer display reference

value .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

2. Press

MARKER ! REFERENCE to change the reference value to the value of the active marker .

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 69)
Fig. 25

Figure 2-17. Example of Setting the Reference V alue Using a Marker

2-20 Making Measurements

Page 70 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 70 · click to enlarge
Page 70

Setting the Electrical Delay

This feature adds phase delay to a variation in phase versus frequency , therefore it is only

applicable for ratioed inputs .

NNNNNNNNNNNNNNNNN

F ORMA T

PHASE .

2. Press

and turn the front panel knob or enter a value from the front panel

MARKER F CTN

keypad to position the marker at a point of interest.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER !DELAY to automatically add or subtract enough line length to the receiver

input to compensate for the phase slope at the active marker position. This e ectively

attens the phase trace around the active marker . Y ou can use this to measure the electrical

length or deviation from linear phase .

A dditional electrical delay adjustments are required on devices without constant group delay

over the measured frequency span.

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 70)
Fig. 26

Figure 2-18. Example of Setting the Electrical Delay Using a Marker

Setting the CW Frequency

1. T o place a marker at the desired CW frequency , press:

MARKER

and either turn the front panel knob or enter the value , followed by

FRE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNN

SEQ

SPECIAL FUNCTIONS

MKR ! CW .

Y ou can use this function to set the marker to a gain peak in an ampli er . After pressing

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MKR ! CW FREQ , activate a CW frequency power sweep to look at the gain compression with

increasing input power .

Show 7 more labels / short text on this page
Press · · Press · · Press · · Making Measurements 2-21
Page 71 of the Agilent Technologies 8753D user manual — N
Page 71 · click to enlarge
Page 71

T o Search for a Speci c Amplitude

These functions place the marker at an amplitude-related point on the trace . If you switch on

tracking, the analyzer searches every new trace for the target point.

Searching for the Maximum Amplitude

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER F CTN

MKR SEARCH to access the marker search menu.

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEARCHMAX to move the active marker to the maximum point on the measurement

trace .

Agilent Technologies 8753D — N diagram (page 71)
Fig. 27

Figure 2-19.

Example of Searching for the Maximum Amplitude Using a Marker

Show 4 more labels / short text on this page
Press · · Press · 2-22 Making Measurements
Page 72 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 72 · click to enlarge
Page 72

Searching for the Minimum Amplitude

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER F CTN

MKR SEARCH to access the marker search menu.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEARCHMIN to move the active marker to the minimum point on the measurement

trace .

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 72)
Fig. 28

Figure 2-20.

Example of Searching for the Minimum Amplitude Using a Marker

Show 4 more labels / short text on this page
Press · · Press · Making Measurements 2-23
Page 73 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 73 · click to enlarge
Page 73

Searching for a T arget Amplitude

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER F CTN

MKR SEARCH to access the marker search menu.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEARCHTARGET to move the active marker to the target point on the measurement

trace .

NNNNNNNNNNNNNNNNNNNN

3. If you want to change the target amplitude value (default is 3 dB), press

enter the new value from the front panel keypad.

4. If you want to search for multiple responses at the target amplitude value , press

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEARCH RIGHT .

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 73)
Fig. 29

Figure 2-21.

Example of Searching for a T arget Amplitude Using a Marker

Show 6 more labels / short text on this page
Press · · Press · TARGET and · SEARCH LEFT and · 2-24 Making Measurements
Page 74 of the Agilent Technologies 8753D user manual — N
Page 74 · click to enlarge
Page 74

Searching for a Bandwidth

The analyzer can automatically calculate and display the 3 dB bandwidth (BW:), center

frequency (CENT:), Q, and loss of the device under test at the center frequency . (Q stands for

\quality factor ," de ned as the ratio of a circuit's resonant frequency to its bandwidth.) These

values are shown in the marker data readout.

1. Press

MARKER

and turn the front panel knob or enter a value from the front panel keypad

to place the marker at the center of the lter passband.

NNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MKR ZERO

MARKER F CTN

MKR SEARCH to access the marker search menu.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

WIDTHS ON to calculate the center stimulus value , bandwidth, and the Q of a

bandpass or band reject shape on the measurement trace .

4. If you want to change the amplitude value (default is 3 dB) that de nes the passband or

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

WIDTH VALUE and enter the new value from the front panel keypad.

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 74)
Fig. 30

Figure 2-22. Example of Searching for a Bandwidth Using Markers

Tracking the Amplitude that Y ou are Searching

1. Set up an amplitude search by following one of the previous procedures in \T o Search for a

Speci c Amplitude ."

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER F CTN

MKR SEARCH

TRACKING ON to track the speci ed amplitude search

with every new trace and put the active marker on that point.

When tracking is not activated, the analyzer nds the speci ed amplitude on the current

sweep and the marker remains at same stimulus value , regardless of changes in the trace

response value with subsequent sweeps .

Show 7 more labels / short text on this page
Press · · Press · rejectband, press · Press · · Making Measurements 2-25
Page 75 of the Agilent Technologies 8753D user manual — MARKER F CTN
Page 75 · click to enlarge
Page 75

T o Calculate the Statistics of the Measurement Data

This function calculates the mean, standard deviation, and peak-to-peak values of the section

of the displayed trace between the active marker and the delta reference . If there is no delta

reference , the analyzer calculates the statistics for the entire trace .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNN

MARKER

MODE MENU

REF=1 to make marker 1 a reference marker .

2. Move marker 1 to any point that you want to reference:

Turn the front panel knob .

OR

Enter the frequency value on the numeric keypad.

NNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER 2 and move marker 2 to any position that you want to measure in reference

to marker 1.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNN

4. Press

MARKER F CTN

MKR MODE MENU

STATS ON to calculate and view the mean, standard

deviation, and peak-to-peak values of the section of the measurement data between the

active marker and the delta reference marker .

An application for this feature is to nd the peak-to-peak value of passband ripple without

searching separately for the maximum and minimum values .

If you are viewing a measurement in the polar or Smith Chart format, the analyzer

calculates the statistics using the rst value of the complex pair (magnitude , real part,

resistance , or conductance).

Agilent Technologies 8753D — MARKER F CTN diagram (page 75)
Fig. 31

Figure 2-23. Example Statistics of Measurement Data

Show 5 more labels / short text on this page
Press · · Press · · 2-26 Making Measurements
Page 76 of the Agilent Technologies 8753D user manual — RESPONSE
Page 76 · click to enlarge
Page 76

Measuring Magnitude and Insertion Phase Response

The analyzer allows you to make two di erent measurements simultaneously . Y ou can make

these measurements in di erent formats for the same parameter . F or example , you could

measure both the magnitude and phase of transmission. Y ou could also measure two di erent

).

This measurement example shows you how to measure the maximum amplitude of a SA W lter

and then how to view the measurement data in the phase format, which provides information

about the phase response .

Measuring the Magnitude Response

1. Connect your test device as shown in Figure 2-24 .
Agilent Technologies 8753D — Chapter 2 Making Measurements diagram (page 76)
Fig. 32 · Chapter 2 Making Measurements

Figure 2-24. Device Connections for Measuring a Magnitude Response

2. Press

PRESET

and choose the measurement settings . F or this example the measurement

parameters are set as follows:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MEAS

TransFWD S21 (B/R)

CENTER

M/

SP AN

M/

NNNNNNNNNNNNNNNNN

MENU

POWER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SCALE REF

AUTO SCALE

CHAN 2

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MEAS

TransFWD S21 (B/R)

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SCALE REF

AUTO SCALE

Y ou may also want to select settings for the number of data points , averaging, and IF

bandwidth.

3. Substitute a thru for the device and perform a response calibration for both channel 1 and

channel 2.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNN

CAL

CALIBRATE MENU

RESPONSE

THRU .

CHAN 1

N

NNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNN

THRU .

RESPONSE

Show 28 more labels / short text on this page
parameters (S · and S · · · · · 134 · · · · 50 · · · · 3 · · · · · · · · · · Press · · Press · Making Measurements 2-27
Page 77 of the Agilent Technologies 8753D user manual — CHAN 2
Page 77 · click to enlarge
Page 77
4. Reconnect your test device .
5. T o better view the measurement trace , press:

SCALE REF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AUTO SCALE

6. T o locate the maximum amplitude of the device response , as shown in Figure 2-25, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER F CTN

MKR SEARCH

SEARCH: MAX

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 77)
Fig. 33

Figure 2-25. Example Magnitude Response Measurement Results

Measuring Insertion Phase Response

7. T o view both the magnitude and phase response of the device , as shown in Figure 2-26,

press:

CHAN 2

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DISPLA Y

DUAL CHAN ON

N

NNNNNNNNNNNNNNNN

PHASE

F ORMA T

The channel 2 portion of Figure 2-26 shows the insertion phase response of the device under

test. The analyzer measures and displays phase over the range of 180

. As phase

changes beyond these values , a sharp 360

transition occurs in the displayed data.

Agilent Technologies 8753D — CHAN 2  diagram (page 77)
Fig. 34

Figure 2-26. Example Insertion Phase Response Measurement

Show 14 more labels / short text on this page
· · · · · · · · · · to +180 · · · 2-28 Making Measurements
Page 78 of the Agilent Technologies 8753D user manual — Chapter 2 Making Measurements
Page 78 · click to enlarge
Page 78

The phase response shown in Figure 2-27 is undersampled; that is , there is more than

phase delay between frequency points . If the = >180

, incorrect phase and delay

information may result. Figure 2-27 shows an example of phase samples being with less

and greater than 180

Agilent Technologies 8753D — Chapter 2 Making Measurements diagram (page 78)
Fig. 35 · Chapter 2 Making Measurements

Figure 2-27. Phase Samples

Undersampling may arise when measuring devices with long electrical length. T o correct this

problem, the frequency span should be reduced, or the number of points increased until is

per point. Electrical delay may also be used to compensate for this e ect (as

shown in the next example procedure).

Show 8 more labels / short text on this page
· · than 180 · · · less than 180 · · Making Measurements 2-29
Page 79 of the Agilent Technologies 8753D user manual — CENTER
Page 79 · click to enlarge
Page 79

Measuring Electrical Length and Phase Distortion

The analyzer mathematically implements a function similar to the mechanical \line stretchers"

of earlier analyzers . This feature simulates a variable length lossless transmission line , which

you can add to or remove from the analyzer's receiver input to compensate for interconnecting

cables , etc . In this example , the electronic line stretcher measures the electrical length of a

SA W lter .

The analyzer allows you to measure the linearity of the phase shift through a device over a

range of frequencies and the analyzer can express it in two di erent ways:

deviation from linear phase

1. Connect your test device as shown in Figure 2-28.

Figure 2-28. Device Connections for Measuring Electrical Length

2. Press

and choose the measurement settings . F or this example , the measurement

PRESET

settings include reducing the frequency span to eliminate under sampled phase response .

Press the following keys as shown:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MEAS

TransFWD S21 (B/R)

CENTER

M/

SP AN

M/

NNNNNNNNNNNNNNNNN

MENU

POWER

NNNNNNNNNNNNNNNNN

F ORMA T

PHASE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SCALE REF

AUTO SCALE

Y ou may also want to select settings for the number of data points , averaging, and IF

bandwidth.

Show 23 more labels / short text on this page
Electrical Length · Phase Distortion · group delay · Measuring Electrical Length · · · · · 134 · · · · 2 · · · · 5 · · · · · · 2-30 Making Measurements
Page 80 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 80 · click to enlarge
Page 80
3. Substitute a thru for the device and perform a response calibration by pressing:

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNN

CAL

CALIBRATE MENU

RESPONSE

THRU

4. Reconnect your test device .
5. T o better view the measurement trace , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AUTO SCALE

SCALE REF

a 2 MHz span. Other lters may require a wider frequency span to see the e ects of phase

shift.

The linearly changing phase is due to the device's electrical length. Y ou can measure this

changing phase by adding electrical length (electrical delay) to compensate for it.

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 80)
Fig. 36

Figure 2-29. Linearly Changing Phase

6. T o place a marker at the center of the band, press:

and turn the front panel knob or enter a value from the front panel keypad.

MARKER

7. T o activate the electrical delay function, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER F CTN

MARKER ! DELAY

This function calculates and adds in the appropriate electrical delay by taking a 10%

span about the marker , measuring the , and computing the delay as the negative of

/frequency .

Show 8 more labels / short text on this page
· · · · · · · Making Measurements 2-31
Page 81 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 81 · click to enlarge
Page 81

N

ELECTRICAL DELAY and turn the front panel knob to increase the

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SCALE REF

electrical length until you achieve the best at line , as shown in Figure 2-30.

The measurement value that the analyzer displays represents the electrical length of your

device relative to the speed of light in free space . The physical length of your device is

related to this value by the propagation velocity of its medium.

V elocity factor is the ratio of the velocity of wave propagation in a coaxial

cable to the velocity of wave propagation in free space . Most cables have a

relative velocity of about 0.66 the speed in free space . This velocity depends

on the relative permittivity of the cable dielectric (

V el ocity F actor =

is the relative permittivity of the cable dielectric

Y ou could change the velocity factor to compensate for propagation velocity by

NNNNNNNNNNNNNN

MORE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

VELOCITY FACTOR (enter the value)

. This will help

CAL

the analyzer to accurately calculate the equivalent distance that corresponds to

the entered electrical delay .

Agilent Technologies 8753D — CAL diagram (page 81)
Fig. 37

Figure 2-30. Example Best Flat Line with A dded Electrical Delay

9. T o display the electrical length, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SCALE REF

ELECTRICAL DELAY

In this example , there is a large amount of electrical delay due to the long electrical length

of the SA W lter under test.

This portion of the example shows you how to measure the linearity of the phase shift over

a range of frequencies . The analyzer allows you to measure this linearity and read it in two

di erent ways: deviation from linear phase , or group delay .

Deviation From Linear Phase

By adding electrical length to \ atten out" the phase response , you have removed the linear

phase shift through your device . The deviation from linear phase shift through your device is

all that remains .

Show 15 more labels / short text on this page
Press · · ) as · r · p · " · r · where · r · pressing · · · · Measuring Phase Distortion · 2-32 Making Measurements
Page 82 of the Agilent Technologies 8753D user manual — F ORMA T
Page 82 · click to enlarge
Page 82
1. F ollow the procedure in \Measuring Electrical Length."
2. T o increase the scale resolution, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SCALE REF

SCALE DIV and turn the front panel knob or enter a value from the front panel

keypad.

3. T o use the marker statistics to measure the maximum peak-to-peak deviation from linear

phase , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER F CTN

MKR MODE MENU

STATS ON

4. A ctivate and adjust the electrical delay to obtain a minimum peak-to-peak value .

It is possible to use delta markers to measure peak-to-peak deviation in only

one portion of the trace , see \T o Calculate the Statistics of the Measurement

Data" located earlier in this Chapter .

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 82)
Fig. 38

Figure 2-31. Deviation From Linear Phase Example Measurement

The phase linearity of many devices is speci ed in terms of group or envelope delay .

The analyzer can translate this information into a related parameter , group delay . Group

delay is the transmission time through your device under test as a function of frequency .

Mathematically , it is the derivative of the phase response which can be approximated by the

following ratio:

/(360 * F)

where is the di erence in phase at two frequencies separated by F . The quantity F is

commonly called the \aperture" of the measurement. The analyzer calculates group delay from

its phase response measurements .

The default aperture is the total frequency span divided by the number of points across the

display (i.e . 201 points or 0.5% of the total span in this example).

1. Continue with the same instrument settings and measurements as in the previous procedure ,

\Deviation from Linear Phase ."

2. T o view the measurement in delay format, as shown in Figure 2-32, press:

F ORMA T

NNNNNNNNNNNNNNNNN

DELAY

SCALE REF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SCALE DIV

Show 11 more labels / short text on this page
· · · · Group Delay · · · · * · * · Making Measurements 2-33
Page 83 of the Agilent Technologies 8753D user manual — A VG
Page 83 · click to enlarge
Page 83
3. T o activate a marker to measure the group delay at a particular frequency , press:

MARKER

and turn the front panel knob or enter a value from the front panel keypad.

Agilent Technologies 8753D — MARKER diagram (page 83)
Fig. 39

Figure 2-32. Group Delay Example Measurement

Group delay measurements may require a speci c aperture (f) or frequency spacing between

measurement points . The phase shift between two adjacent frequency points must be less than

, otherwise incorrect group delay information may result.

4. T o vary the e ective group delay aperture from minimum aperture (no smoothing) to

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

approximately 1% of the frequency span, press:

A VG

SMOOTHING ON .

When you increase the aperture , the analyzer removes ne grain variations from the

response . It is critical that you specify the group delay aperture when you compare group

delay measurements .

Agilent Technologies 8753D — A VG diagram (page 83)
Fig. 40

Figure 2-33. Group Delay Example Measurement with Smoothing

Show 4 more labels / short text on this page
· · · 2-34 Making Measurements
Page 84 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 84 · click to enlarge
Page 84
5. T o increase the e ective group delay aperture , by increasing the number of measurement

points over which the analyzer calculates the group delay , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SMOOTHING APERTURE

As the aperture is increased the \smoothness" of the trace improves markedly , but at the

expense of measurement detail.

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 84)
Fig. 41

Figure 2-34.

Group Delay Example Measurement with Smoothing Aperture Increased

5 · · Making Measurements 2-35

Page 85 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 85 · click to enlarge
Page 85

T esting A Device with Limit Lines

Limit testing is a measurement technique that compares measurement data to constraints that

you de ne . Depending on the results of this comparison, the analyzer will indicate if your

device either passes or fails the test.

Limit testing is implemented by creating individual at, sloping, and single point limit lines on

the analyzer display . When combined, these lines can represent the performance parameters

for your device under test. The limit lines created on each measurement channel are

independent of each other .

This example measurement shows you how to test a bandpass lter using the following

procedures:

creating at limit lines

creating sloping limit lines

creating single point limit lines

running a limit test

Setting Up the Measurement P arameters

1. Connect your test device as shown in Figure 2-35.

Figure 2-35. Connections for S A W Filter Example Measurement

2. Press

and choose the measurement settings . F or this example the measurement

PRESET

settings are as follows:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MEAS

TransFWD S21 (B/R)

CENTER

M/

SP AN

M/

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SCALE REF

AUTO SCALE

Y ou may also want to select settings for the number of data points , power , averaging, and IF

bandwidth.

3. Substitute a thru for the device and perform a response calibration by pressing:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNN

CAL

CALIBRATE MENU

RESPONSE

THRU

Show 16 more labels / short text on this page
editing limit segments · · · · · 134 · · · · 50 · · · · · · 2-36 Making Measurements
Page 86 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 86 · click to enlarge
Page 86
4. Reconnect your test device .
5. T o better view the measurement trace , press:

SCALE REF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AUTO SCALE

Creating Flat Limit Lines

In this example procedure , the following at limit line values are set:

Frequency Range : P ower Range

127 MHz to 140 MHz : 27 dB to 21 dB

100 MHz to 123 MHz : 200 dB to 65 dB

146 MHz to 160 MHz : 200 dB to 65 dB

The minimum value for measured data is 200 dB .

1. T o access the limits menu and activate the limit lines , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LIMIT MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LIMIT LINE ON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EDIT LIMIT LINE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CLEAR LIST

NNNNNNNNNNN

YES

SYSTEM

2. T o create a new limit line , press:

NNNNNNNNNNN

ADD

The analyzer generates a new segment that appears on the center of the display .

3. T o specify the limit's stimulus value , test limits (upper and lower), and the limit type , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STIMULUS VALUE

M/

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

UPPER LIMIT

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOWER LIMIT

NNNNNNNNNNNNNN

DONE

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Y ou could also set the upper and lower limits by using the

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DELTA LIMITS keys . T o use these keys for the entry , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MIDDLE VALUE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DELTA LIMITS

This would correspond to a test speci cation of 24 3 dB .

4. T o de ne the limit as a at line , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LIMIT TYPE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FLAT LINE

NNNNNNNNNNNNNNNNNNNN

RETURN

Show 19 more labels / short text on this page
· · · · 127 · · 21 · · · 27 · · · MIDDLE VALUE and · 24 · · · 3 · · Making Measurements 2-37
Page 87 of the Agilent Technologies 8753D user manual — NNNNNNNNNNN
Page 87 · click to enlarge
Page 87
5. T o terminate the at line segment by establishing a single point limit, press:

NNNNNNNNNNN

ADD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STIMULUS VALUE

M/

NNNNNNNNNNNNNN

DONE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LIMIT TYPE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SINGLE POINT

NNNNNNNNNNNNNNNNNNNN

RETURN

Figure 2-36 shows the at limit lines that you have just created with the following

parameters:

stimulus from 127 MHz to 140 MHz

upper limit of 21 dB

lower limit of 27 dB

Agilent Technologies 8753D — NNNNNNNNNNN diagram (page 87)
Fig. 42

Figure 2-36. Example Flat Limit Line

6. T o create a limit line that tests the low side of the lter , press:

NNNNNNNNNNN

ADD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STIMULUS VALUE

M/

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

UPPER LIMIT

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOWER LIMIT

NNNNNNNNNNNNNN

DONE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNN

LIMIT TYPE

FLAT LINE

RETURN

NNNNNNNNNNN

ADD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STIMULUS VALUE

M/

NNNNNNNNNNNNNN

DONE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LIMIT TYPE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SINGLE POINT

NNNNNNNNNNNNNNNNNNNN

RETURN

Show 13 more labels / short text on this page
140 · · 100 · · 65 · · · 200 · · · 123 · · 2-38 Making Measurements
Page 88 of the Agilent Technologies 8753D user manual — NNNNNNNNNNN
Page 88 · click to enlarge
Page 88
7. T o create a limit line that tests the high side of the bandpass lter , press:

NNNNNNNNNNN

ADD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STIMULUS VALUE

M/

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

UPPER LIMIT

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOWER LIMIT

NNNNNNNNNNNNNN

DONE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNN

LIMIT TYPE

FLAT LINE

RETURN

NNNNNNNNNNN

ADD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STIMULUS VALUE

M/

NNNNNNNNNNNNNN

DONE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNN

LIMIT TYPE

SINGLE POINT

RETURN

Agilent Technologies 8753D — NNNNNNNNNNN diagram (page 88)
Fig. 43

Figure 2-37. Example Flat Limit Lines

Creating a Sloping Limit Line

This example procedure shows you how to make limits that test the shape factor of a SA W

lter . The following limits are set:

Frequency Range : P ower Range

123 MHz to 125 MHz : 65 dB to 26 dB

144 MHz to 146 MHz : 26 dB to 65 dB

1. T o access the limits menu and activate the limit lines , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNN

SYSTEM

LIMIT MENU

LIMIT LINE ON

EDIT LIMIT LINE

CLEAR LIST

YES

2. T o establish the start frequency and limits for a sloping limit line that tests the low side of

the lter , press:

NNNNNNNNNNN

ADD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STIMULUS VALUE

M/

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

UPPER LIMIT

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOWER LIMIT

NNNNNNNNNNNNNN

DONE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LIMIT TYPE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SLOPING LINE

NNNNNNNNNNNNNNNNNNNN

RETURN

Show 21 more labels / short text on this page
146 · · 65 · · · 200 · · · 160 · · · · 123 · · 65 · · · 200 · · · Making Measurements 2-39
Page 89 of the Agilent Technologies 8753D user manual — NNNNNNNNNNN
Page 89 · click to enlarge
Page 89
3. T o terminate the lines and create a sloping limit line , press:

NNNNNNNNNNN

ADD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STIMULUS VALUE

M/

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

UPPER LIMIT

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOWER LIMIT

NNNNNNNNNNNNNN

DONE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNN

LIMIT TYPE

SINGLE POINT

RETURN

4. T o establish the start frequency and limits for a sloping limit line that tests the high side of

the lter , press:

NNNNNNNNNNN

ADD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STIMULUS VALUE

M/

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

UPPER LIMIT

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOWER LIMIT

NNNNNNNNNNNNNN

DONE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNN

LIMIT TYPE

SLOPING LINE

RETURN

5. T o terminate the lines and create a sloping limit line , press:

NNNNNNNNNNN

ADD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STIMULUS VALUE

M/

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

UPPER LIMIT

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOWER LIMIT

NNNNNNNNNNNNNN

DONE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LIMIT TYPE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SINGLE POINT

NNNNNNNNNNNNNNNNNNNN

RETURN

Y ou could use this type of limit to test the shape factor of a lter .

Agilent Technologies 8753D — NNNNNNNNNNN diagram (page 89)
Fig. 44

Figure 2-38. Sloping Limit Lines

Show 25 more labels / short text on this page
125 · · 26 · · · 200 · · · 144 · · 26 · · · 200 · · · 146 · · 65 · · · 200 · · · 2-40 Making Measurements
Page 90 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNN
Page 90 · click to enlarge
Page 90

Creating Single P oint Limits

In this example procedure , the following limits are set:

from 23 dB to 28.5 dB at 141 MHz

from 23 dB to 28.5 dB at 126.5 MHz

1. T o access the limits menu and activate the limit lines , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNN

SYSTEM

LIMIT MENU

LIMIT LINE ON

EDIT LIMIT LINE

CLEAR LIST

YES

2. T o designate a single point limit line , as shown in Figure 2-39, you must de ne two pointers;

downward pointing, indicating the upper test limit

upward pointing, indicating the lower test limit

Press:

N

ADD

NNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STIMULUS VALUE

M/

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

UPPER LIMIT

N

LOWER LIMIT

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

28.5

N

DONE

NNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LIMIT TYPE

SINGLE POINT

NNNNNNNNNNNNNNNNNNNN

RETURN

NNNNNNNNNNN

ADD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STIMULUS VALUE

126.5

M/

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

UPPER LIMIT

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOWER LIMIT

28.5

NNNNNNNNNNNNNN

DONE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LIMIT TYPE

SINGLE POINT

NNNNNNNNNNNNNNNNNNNN

RETURN

Agilent Technologies 8753D — NNNNNNNNNNNNNN diagram (page 90)
Fig. 45

Figure 2-39. Example Single P oints Limit Line

Show 16 more labels / short text on this page
· · 141 · · 23 · · · · · · 23 · · · · · Making Measurements 2-41
Page 91 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNN
Page 91 · click to enlarge
Page 91

This example shows you how to edit the upper limit of a limit line .

1. T o access the limits menu and activate the limit lines , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SYSTEM

LIMIT MENU

LIMIT LINE ON

EDIT LIMIT LINE

2. T o move the pointer symbol ( > ) on the analyzer display to the segment you wish to modify ,

press:

N

SEGMENT

NNNNNNNNNNNNNNNNNNNNNN

* or

OR

NNNNNNNNNNNNNNNNNNNNNNN

SEGMENT and enter the segment number followed by

3. T o change the upper limit (for example , 20) of a limit line , press:

NNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNN

EDIT

UPPER LIMIT

DONE

1. T o access the limits menu and activate the limit lines , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SYSTEM

LIMIT MENU

LIMIT LINE ON

EDIT LIMIT LINE

2. T o move the pointer symbol ( > ) on the analyzer display to the segment you wish to delete ,

press:

NNNNNNNNNNNNNNNNNNNNNNN

SEGMENT

* or

OR

NNNNNNNNNNNNNNNNNNNNNNN

SEGMENT and enter the segment number followed by

3. T o delete the segment that you have selected with the pointer symbol, press:

NNNNNNNNNNNNNNNNNNNN

DELETE

Show 15 more labels / short text on this page
Editing Limit Segments · · · + · repeatedly · · · · Deleting Limit Segments · · · + · repeatedly · · 2-42 Making Measurements
Page 92 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 92 · click to enlarge
Page 92

Running a Limit T est

1. T o access the limits menu and activate the limit lines , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SYSTEM

LIMIT MENU

LIMIT LINE ON

EDIT LIMIT LINE

Reviewing the Limit Line Segments

The limit table data that you have previously entered is shown on the analyzer display .

2. T o verify that each segment in your limits table is correct, review the entries by pressing:

NNNNNNNNNNNNNNNNNNNNNNN

SEGMENT

* and

3. T o modify an incorrect entry , refer to the \Editing Limit Segments" procedure , located

earlier in this section.

A ctivating the Limit T est

4. T o activate the limit test and the beep fail indicator , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SYSTEM

LIMIT MENU

LIMIT TEST ON

BEEP FAIL ON

Selecting the beep fail indicator

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

BEEP FAIL ON is optional and will add

approximately 50 ms of sweep cycle time . Because the limit test will still work

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

if the limits lines are o , selecting

LIMIT LINE ON is also optional.

The limit test results appear on the right side on the analyzer display . The analyzer indicates

whether the lter passes or fails the de ned limit test:

The message FAIL will appear on the right side of the display if the limit test fails .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

The analyzer beeps if the limit test fails and if

BEEP FAIL ON has been selected.

The analyzer alternates a red trace where the measurement trace is out of limits .

A TTL signal on the rear panel BNC connector \LIMIT TEST" provides a pass/fail (5 V/0 V)

indication of the limit test results .

Show 7 more labels / short text on this page
· · + · · · · Making Measurements 2-43
Page 93 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 93 · click to enlarge
Page 93

O setting Limit Lines

The limit o set functions allow you to adjust the limit lines to the frequency and output level

of your device . F or example , you could apply the stimulus o set feature for testing tunable

lters . Or , you could apply the amplitude o set feature for testing variable attenuators , or

passband ripple in lters with variable loss .

This example shows you the o set feature and the limit test failure indications that can appear

on the analyzer display .

1. T o o set all of the segments in the limit table by a xed frequency , (for example , 3 MHz),

press:

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LIMIT MENU

LIMIT LINE OFFSETS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STIMULUS OFFSET

M/

The analyzer beeps and a FAIL notation appears on the analyzer display , as shown in

Figure 2-40.

Agilent Technologies 8753D — SYSTEM diagram (page 93)
Fig. 46

Figure 2-40. Example Stimulus O set of Limit Lines

2. T o return to 0 Hz o set, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STIMULUS OFFSET

3. T o o set all of the segments in the limit table by a xed amplitude , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AMPLITUDE OFFSET

The analyzer beeps and a FAIL notation appears on the analyzer display .

4. T o return to 0 dB o set, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AMPLITUDE OFFSET

Show 11 more labels / short text on this page
· · 3 · · 0 · · 5 · · 0 · · 2-44 Making Measurements
Page 94 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 94 · click to enlarge
Page 94

Gain compression occurs when the input power of an ampli er is increased to a level that

reduces the gain of the ampli er and causes a nonlinear increase in output power . The point at

which the gain is reduced by 1 dB is called the 1 dB compression point. The gain compression

will vary with frequency , so it is necessary to nd the worst case point of gain compression in

the frequency band.

Once that point is identi ed, you can perform a power sweep of that CW frequency to measure

the input power at which the 1 dB compression occurs and the absolute power out (in dBm) at

compression. The following steps provide detailed instruction on how to apply various features

of the analyzer to accomplish these measurements .

Agilent Technologies 8753D — Chapter 2 Making Measurements diagram (page 94)
Fig. 47 · Chapter 2 Making Measurements

Figure 2-41. Diagram of Gain Compression

1. Set up the stimulus and response parameters for your ampli er under test. T o reduce the

e ect of noise on the trace , press:

A V G

NNNNNNNNNNNNNNNNN

IF BW

NNNNNNNNNNNNNN

2. P erform the desired error correction procedure . Refer to Chapter 5, \Optimizing

Measurement Results ," for instructions on how to make a measurement correction.

3. Hook up the ampli er under test.
4. T o produce a normalized trace that represents gain compression, perform either step 5

or step 6. (Step 5 uses trace math and step 6 uses uncoupled channels and the display

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

D1/D2 to D2 ON .)

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNN

DISPLA Y

DATA ! MEMORY

DATA/MEM to produce a normalized trace .

6. T o produce a normalized trace , perform the following steps:

a.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DISPLA Y

DUAL CHANNEL ON to view both channels simultaneously .

Show 12 more labels / short text on this page
Measuring Gain Compression · · · 1000 · 1 · · function · Press · · Press · and select · Making Measurements 2-45
Page 95 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 95 · click to enlarge
Page 95

b. T o uncouple the channel stimulus so that the channel power will be uncoupled, press:

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

COUPLED CH OFF

This will allow you to separately increase the power for channel 2 and channel 1 ,

so that you can observe the gain compression on channel 2 while channel 1 remains

unchanged.

c. T o display the ratio of channel 2 data to channel 1 data on the channel 2 display , press:

CHAN 2

NNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DISPLAY

MORE

D2/D1 to D2 ON

This produces a trace that represents gain compression only .

NNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER

MARKER 1 and position the marker at approximately mid-span.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SCALE REF

SCALE/DIV

to change the scale to 1 dB per division.

NNNNNNNNNNNNNNNNN

MENU

POWER .

10. Increase the power until you observe approximately 1 dB of compression on channel 2 ,

using the step keys or the front panel knob .

11. T o locate the worst case point on the trace , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER F CTN

MKR SEARCH

SEARCH:MIN

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 95)
Fig. 48

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Figure 2-42. Gain Compression Using Linear Sweep and

D2/D1 to D2 ON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

COUPLED CH OFF was selected, recouple the channel stimulus by pressing:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MENU

COUPLED CH ON

13. T o place the marker exactly on a measurement point, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER MODE MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKERS:DISCRETE

MARKER F CTN

Show 18 more labels / short text on this page
· · · Press · · Press · · 1 · Press · · · · If · · · · · 2-46 Making Measurements
Page 96 of the Agilent Technologies 8753D user manual — N
Page 96 · click to enlarge
Page 96
14. T o set the CW frequency before going into the power sweep mode , press:

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SPECIAL FUNCTIONS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER ! CW

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MENU

SWEEP TYPE MENU

POWER SWEEP .

16. Enter the start and stop power levels for the sweep .

Now channel 1 is displaying a gain compression curve . (Do not pay attention to channel 2

at this time .)

17. T o maintain the calibration for the CW frequency , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL

INTERPOL ON

CORRECTION ON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CHAN 2

DISPLA Y

DUAL CHANNEL ON .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

D2/D1 to D2 ON was selected, press

MORE

D2/D1 to D2 OFF .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNN

MEAS

INPUT PORTS

B .

Now channel 2 displays absolute output power (in dBm) as a function of power input.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SCALE REF

SCALE/DIV

to change the scale of channel 2 to 10 dB per division.

22. Press

CHAN 1

to change the scale of channel 1 to 1 dB per division.

A receiver calibration will improve the accuracy of this measurement. Refer to

Chapter 5, \Optimizing Measurement Results ."

23. Press

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER MODE MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKERS:COUPLED .

MARKER F CTN

24. T o nd the 1 dB compression point on channel 1 , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER F CTN

MKR SEARCH

SEARCH:MAX

NNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER

MKR ZERO

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEARCH:TARGET

MARKER F CTN

MKR SEARCH

25. Press

CHAN 2

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER

MKR MODE MENU

MARKERS:UNCOUPLED .

26. T o take the channel 2 marker out of the mode so that it reads the absolute output power

of the ampli er (in dBm), press:

N

MODE MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MODE OFF

MARKER

Show 30 more labels / short text on this page
· · Press · · · · Press · · If · Press · · Press · · 10 · 1 · · · · · · · · 1 · · 1 · · · · · Making Measurements 2-47
Page 97 of the Agilent Technologies 8753D user manual — Chapter 2 Making Measurements
Page 97 · click to enlarge
Page 97
Agilent Technologies 8753D — Chapter 2 Making Measurements diagram (page 97)
Fig. 49 · Chapter 2 Making Measurements

Figure 2-43. Gain Compression Using P ower Sweep

2-48 Making Measurements

Page 98 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 98 · click to enlarge
Page 98

Measuring Gain and Reverse Isolation Simultaneously

Since an ampli er will have high gain in the forward direction and high isolation in the reverse

direction, the gain (S

) will be much greater than the reverse isolation (S

). Therefore , the

power you apply to the input of the ampli er for the forward measurement (S

considerably lower than the power you apply to the output for the reverse measurement

(S

). By applying low power in the forward direction, you'll prevent the ampli er from being

saturated. A higher power in the reverse direction keeps noise from being a factor in the

measurement and accounts for any losses caused by attenuators or couplers on the ampli er's

output needed to lower the output power into the analyzer . The following steps demonstrate

the features that best accomplish these measurements .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

COUPLED CH ON .

MENU

Coupling the channels allows you to have the same frequency range and calibration applied

to channel 1 and channel 2 .

NNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

2. Press

POWER

PORT POWER [UNCOUPLED] .

Uncoupling the port power allows you to apply di erent power levels at each port. In

Figure 2-44 , the port 1 power is set to 25 dBm for the gain measurement (S21) and the

port 2 power is set to 0 dBm for the reverse isolation measurement (S12).

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNN

CHAN 1

MEAS

TransFWD S21 (B/R)

MENU

POWER and set the power level for

port 1.

CHAN 2

N

TransREV S12 (A/R)

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

N

NNNNNNNNNNNNNNNN

POWER and set the power level for

MEAS

MENU

port 2.

5. P erform an error-correction and connect the ampli er to the network analyzer . Refer to the

\Optimizing Measurement Results" Chapter for error-correction procedures .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DISPLA Y

DUAL CHAN ON .

Y ou can view both measurements simultaneously by using the dual channel display mode .

Refer to Figure 2-44. If the port power levels are in di erent power ranges , one of the

displayed measurements will not be continually updated and the annotation tsH will appear

on the left side of the display . Refer to \Source Attenuator Switch Protection"section in

Chapter 6, \Application and Operation Concepts ," for information on how to override this

state .

Show 12 more labels / short text on this page
) should be · Press · · Press · · · Press · · · Press · · Making Measurements 2-49
Page 99 of the Agilent Technologies 8753D user manual — Chapter 2 Making Measurements
Page 99 · click to enlarge
Page 99

T o obtain best accuracy , you should set the power levels prior to performing

the calibration. However , the analyzer compensates for nominal power changes

you make during a measurement, so that the error correction still remains

approximately valid. In these cases , the Cor annunciator will change to C? .

Agilent Technologies 8753D — Chapter 2 Making Measurements diagram (page 99)
Fig. 50 · Chapter 2 Making Measurements

Figure 2-44. Gain and Reverse Isolation

2-50 Making Measurements

Page 100 of the Agilent Technologies 8753D user manual — SYSTEM
Page 100 · click to enlarge
Page 100

Measurements Using the Tuned Receiver Mode

In the tuned receiver mode , the analyzer's receiver operates independently of any signal

source . This mode is not phase-locked and functions in all sweep types . The analyzer tunes

the receiver to a synthesized CW input signal at a precisely speci ed frequency . All phase

lock routines are bypassed, increasing sweep speed signi cantly . The external source must

be synthesized, and must drive the analyzer's external frequency reference . The analyzer's

internal source frequency is not accurate , and the internal source should not be used in the

tuned receiver mode .

Using the analyzer's tuned receiver mode is useful for automated test applications where an

external synthesized source is available and applications where speed is important. Although

the tuned receiver mode can function in all sweep types , it is typically used in CW applications .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

A ctivate the tuned receiver mode by pressing

SYSTEM

INSTRUMENT MODE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TUNED RECEIVER .

2. Connect the equipment as shown in Figure 2-45 to perform a CW measurement using the

tuned receiver mode .

Agilent Technologies 8753D — SYSTEM diagram (page 100)
Fig. 51

Figure 2-45. Typical T est Setup for Tuned Receiver Mode

Tuned receiver mode in-depth description

30 kHz to 3 GHz (6 GHz for Option 006)

All sweep types may be used.

Show 5 more labels / short text on this page
Typical test setup · · Frequency Range · Compatible Sweep Types · Making Measurements 2-51
Page 101 of the Agilent Technologies 8753D user manual — Chapter 2 Making Measurements
Page 101 · click to enlarge
Page 101

An analyzer in tuned receiver mode can receive input signals into PORT 1, PORT 2, or R

CHANNEL IN.

Input power range speci cations are provided in Chapter 7, \Speci cations and Measurement

Uncertainties ."

External Source Requirements · 2-52 Making Measurements

Page 102 of the Agilent Technologies 8753D user manual — Chapter 2 Making Measurements
Page 102 · click to enlarge
Page 102

T est sequencing allows you to automate repetitive tasks . As you make a measurement, the

analyzer memorizes the keystrokes . Later you can repeat the entire sequence by pressing a

single key . Because the sequence is de ned with normal measurement keystrokes , you do not

need additional programming expertise . Subroutines and limited decision-making increases the

exibility of test sequences . In addition, the GPIO outputs can be controlled in a test sequence ,

and the GPIO inputs can be tested in a sequence for conditional branching. F or in-depth

sequencing information, refer to \T est Sequencing" in Chapter 6, \Application and Operation

Concepts ."

The test sequence function allows you to create , title , save , and execute up to six independent

sequences internally .

Y ou can also save sequences to disk and transfer them between the analyzer and an external

computer controller .

The following procedures are based on an actual measurement example , that shows you how to

do the following:

change a sequence title

name les generated by a sequence

There are also three example sequences:

Show 13 more labels / short text on this page
T est Sequencing · create a sequence · title a sequence · edit a sequence · clear a sequence · store a sequence · load a sequence · purge a sequence · print a sequence · cascading multiple sequences · loop counter sequence · limit test sequence · Making Measurements 2-53
Page 103 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 103 · click to enlarge
Page 103
1. T o enter the sequence creation mode , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

NEW SEQ/MODIFY SEQ

As shown in Figure 2-46, a list of instructions appear on the analyzer display to help you

create or edit a sequence .

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 103)
Fig. 52

Figure 2-46. T est Sequencing Help Instructions

2. T o select a sequence position in which to store your sequence , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQUENCE 1 SEQ1

This choice selects sequence position #1. The default title is SEQ1 for this sequence .

Refer to \Changing the Sequence Title ," (located later in this Chapter) for information on

how to modify a sequence title .

Show 6 more labels / short text on this page
Creating a Sequence · · · d a · c b · 2-54 Making Measurements
Page 104 of the Agilent Technologies 8753D user manual — LOCAL
Page 104 · click to enlarge
Page 104
3. T o create a test sequence , enter the parameters for the measurement that you wish to make .

F or this example , a SA W lter measurement is set up with the following parameters:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SELECT DISK

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INTERNAL MEMORY

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNN

RETURN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECALL STATE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TransFWD S21 (B/R)

MEAS

NNNNNNNNNNNNNNNNNNNNNNN

LOG MAG

F ORMA T

CENTER

M/

SP AN

M/

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNN

SCALE REF

AUTOSCALE

The above keystrokes will create a displayed list as shown:

RECALL PRST STATE

TransFWD S21 (B/R)

LOG MAG

CENTER

SPAN

SCALE/DIV

AUTO SCALE

4. T o complete the sequence creation, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DONE SEQ MODIFY

SEQ

When you create a sequence , the analyzer stores it in volatile memory where

it will be lost if you switch o the instrument power (except for sequence #6

which is stored in the analyzer non-volatile memory). However , you may store

sequences to a oppy disk.

T o run a stored test sequence , press:

and the softkey labeled with desired sequence number

PRESET

or , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

DO SEQUENCE and the softkey labeled with the desired sequence number .

T o stop a sequence before it has nished, press

LOCAL

.

Show 27 more labels / short text on this page
· · · · · · · · 134 · · · · 50 · · · · Start of Sequence · 134 M/u · 50 M/u · · · Running a Sequence · · · · Stopping a Sequence · Making Measurements 2-55
Page 105 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 105 · click to enlarge
Page 105
1. T o enter the creation/editing mode , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

NEW SEQ/MODIFY SEQ

2. T o select the particular test sequence you wish to modify (sequence 1 in this example),

press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQUENCE 1 SEQ1

3. T o move the cursor to the command that you wish to delete , press:

* or

If you use the

* key to move the cursor through the list of commands , the commands are

actually performed when the cursor points to them. This feature allows the sequence to

be tested one command at a time .

If you wish to scroll through the sequence without executing each line as you do so , you

can press the

key and scroll through the command list backwards .

4. T o delete the selected command, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

DONE SEQ MODIFY to exit the modify (edit) mode .

1. T o enter the creation/editing mode , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

NEW SEQ/MODIFY SEQ

2. T o select the particular test sequence you wish to modify (sequence 1 in this example),

press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQUENCE 1 SEQ1

3. T o insert a command, move the cursor to the line immediately above the line where you

want to insert a new command, by pressing:

* or

If you use the

* key to move the cursor through the list of commands , the commands are

actually performed when the cursor points to them. This feature allows the sequence to

be tested one command at a time .

If you wish to scroll through the sequence without executing each line as you do so , you

can press the

key and scroll through the command list backwards .

4. T o enter the new command, press the corresponding analyzer front panel keys . F or example ,

if you want to activate the averaging function, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

A VG

AVERAGING ON

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

DONE SEQ MODIFY to exit the modify (edit) mode .

Show 24 more labels / short text on this page
Editing a Sequence · Deleting Commands · · · · + · · + · · (backspace key) · Press · · Inserting a Command · · · · · + · + · · · Press · · 2-56 Making Measurements
Page 106 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNN
Page 106 · click to enlarge
Page 106
1. T o enter the creation/editing mode , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRESET

SEQ

NEW SEQ/MODIFY SEQ

2. T o select the particular test sequence you wish to modify , (sequence 1 in this example),

press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQUENCE 1 SEQ1

The following list is the commands entered in \Creating a Sequence ." Notice that for longer

sequences , only a portion of the list can appear on the screen at one time .

RECALL PRST STATE

TransFWD S21 (B/R)

LOG MAG

CENTER

SPAN

SCALE/DIV

AUTO SCALE

3. T o change a command (for example , the span value from 50 MHz to 75 MHz) move the

cursor (! ) next to the command that you wish to modify , press:

* or

If you use the

* key to move the cursor through the list of commands , the commands are

actually performed when the cursor points to them. This feature allows the sequence to

be tested one command at a time .

If you wish to scroll through the sequence without executing each line as you do so , you

can press the

key and scroll through the command list backwards .

4. T o delete the current command (for example , span value), press:
5. T o insert a new value (for example , 75 MHz), press:

M/

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

DONE SEQ MODIFY to exit the modify (edit) mode .

Clearing a Sequence from Memory

1. T o enter the menu where you can clear a sequence from memory , press:

NNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

MORE

CLEAR SEQUENCE

2. T o clear a sequence , press the softkey of the particular sequence .
Show 21 more labels / short text on this page
Modifying a Command · · · · Start of Sequence · 134 M/u · 50 M/u · · + · · + · · · · 75 · · Press · · · · Making Measurements 2-57
Page 107 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 107 · click to enlarge
Page 107

Changing the Sequence Title

If you are storing sequences on a disk, you should replace the default titles (SEQ1, SEQ2 . . . ).

1. T o select a sequence that you want to retitle , press:

NNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

MORE

TITLE SEQUENCE and select the particular sequence softkey .

The analyzer shows the available title characters . The current title is displayed in the

upper-left corner of the screen.

2. Y ou can create a new lename in two ways:

If you have an attached DIN keyboard, you can press

and then type the new lename .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

If you do not have an attached DIN keyboard, press

ERASE TITLE and turn the front

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

panel knob to point to the characters of the new lename , pressing

you stop at each character .

The analyzer cannot accept a title ( le name) that is longer than eight characters . Y our titles

must also begin with a letter , and contain only letters and numbers .

N

NNNNNNNNNNNNN

T o complete the titling, press

DONE .

Naming Files Generated by a Sequence

The analyzer can automatically increment the name of a le that is generated by a sequence

using a loop structure . (See example ''Generating Files in a Loop Counter Example Sequence''

later in this chapter .)

T o access the sequence lename menu, press:

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FILE UTILITIES

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQUENCE FILENAMING

This menu presents two choices:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FILE NAME FILE0 supplies a name for the saved state and/or data le . This also brings up

the Title File Menu.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOT NAME PLOTFILE supplies a name for the plot le generated by a plot-to-disk command.

This also brings up the Title File Menu.

The above keys show the current lename in the 2nd line of the softkey .

When titling a le for use in a loop function, you are restricted to only 2 characters in the

lename due to the 6 character length of the loop counter keyword \[LOOP]." When the le is

actually written, the [LOOP] keyword is expanded to only 5 ASCII characters (digits), resulting

in a 7 character lename .

After entering the 2 character lename , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOOP COUNTER

NNNNNNNNNNNNNN

DONE

Show 7 more labels / short text on this page
· · f6 · SELECT LETTER as · · · 2-58 Making Measurements
Page 108 of the Agilent Technologies 8753D user manual — SEQ
Page 108 · click to enlarge
Page 108

Storing a Sequence on a Disk

1. T o format a disk, refer to Chapter 4, \Printing, Plotting, and Saving Measurement Results ."
2. T o save a sequence to the internal disk, press:

NNNNNNNNNNNNNN

MORE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STORE SEQ TO DISK and select the particular sequence softkey .

SEQ

The disk drive access light should turn on brie y . When it goes out, the sequence has been

saved.

The analyzer will overwrite a le on the disk that has the same title .

· · Making Measurements 2-59

Page 109 of the Agilent Technologies 8753D user manual — SEQ
Page 109 · click to enlarge
Page 109

Loading a Sequence from Disk

F or this procedure to work, the desired le must exist on the disk in the analyzer drive .

1. T o view the rst six sequences on the disk, press:

NNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

MORE

LOAD SEQ FROM DISK

READ SEQ FILE TITLS

If the desired sequence is not among the rst six les , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

READ SEQ FILE TITLS until the desired le name appears .

2. Press the softkey next to the title of the desired sequence . The disk access light should

illuminate brie y .

If you know the title of the desired sequence , you can title the sequence (1-6)

with the name , and load the sequence . This is also how you can control the

sequence number of an imported titled sequence .

Purging a Sequence from Disk

1. T o view the contents of the disk (six titles at a time), press:

NNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

MORE

STORE SEQ TO DISK

PURGE SEQUENCES

READ SEQ FILE TITLS

If the desired sequence is not among the rst six les , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

READ SEQ FILE TITLS until the desired le name appears .

2. Press the softkey next to the title of the desired sequence . The disk access light should

illuminate brie y .

1. Con gure a compatible printer to the analyzer . (Refer to Chapter 11, \Compatible

P eripherals .")

2. T o print a sequence , press:

SEQ

NNNNNNNNNNNNNN

MORE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRINT SEQUENCE and the softkey for the desired sequence .

If the sequence is on a disk, load the sequence (as described in a previous

procedure) and then follow the printing sequence .

Show 8 more labels / short text on this page
· · · · Printing a Sequence · · · 2-60 Making Measurements
Page 110 of the Agilent Technologies 8753D user manual — PRESET
Page 110 · click to enlarge
Page 110

Cascading Multiple Example Sequences

By cascading test sequences , you can create subprograms for a larger test sequence . Y ou can

also cascade sequences to extend the length of test sequences to greater than 200 lines .

In this example , you are shown two sequences that have been cascaded. Y ou can do this by

having the last command in sequence 1 call sequence position 2, regardless of the sequence

title . Because sequences are identi ed by position, not title , the call operation will always go to

the sequence loaded into the given position.

1. T o create the example multiple sequences , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

NEW SEQ/MODIFY SEQ

SEQUENCE 1 SEQ1

CENTER

M/

SP AN

M/

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DO SEQUENCE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQUENCE 2

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DONE SEQ MODIFY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NEW SEQ/MODIFY SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQUENCE 2 SEQ2

N

MEAS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TransFWD S21 (B/R)

NNNNNNNNNNNNNNNNNNNNNNN

F ORMA T

LOG MAG

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SCALE REF

AUTOSCALE

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

DONE SEQ MODIFY

The following sequences will be created:

SEQUENCE SEQ1

CENTER

SPAN

DO SEQUENCE

SEQUENCE 2

SEQUENCE SEQ2

TransFWD S21 (B/R)

LOG MAG

SCALE/DIV

AUTO SCALE

Y ou can extend this process of calling the next sequence from the last line of the present

sequence to 6 internal sequences , or an unlimited number of externally stored sequences .

2. T o run both sequences , press:

PRESET

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQUENCE 1 SEQ1

Show 27 more labels / short text on this page
· · · · 134 · · · · 50 · · · · · · · · · · · · Start of Sequence · 134 M/u · 50 M/u · Start of Sequence · · · Making Measurements 2-61
Page 111 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 111 · click to enlarge
Page 111

Loop Counter Example Sequence

This example shows you the basic steps necessary for constructing a looping structure within

a test sequence . A typical application of this loop counter structure is for repeating a speci c

measurement as you step through a series of CW frequencies or dc bias levels . F or an example

application, see \Fixed IF Mixer Measurements" in Chapter 3.

1. T o create a sequence that will set the initial value of the loop counter , and call the sequence

that you want to repeat, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

NEW SEQ/MODIFY SEQ

SEQUENCE 1 SEQ1

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SPECIAL FUNCTIONS

DECISION MAKING

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOOP COUNTER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

DO SEQUENCE

SEQUENCE 2

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DONE SEQ MODIFY

This will create a displayed list as shown:

SEQUENCE LOOP 1

LOOP COUNTER

DO SEQUENCE

SEQUENCE 2

T o create a second sequence that will perform a desired measurement function,

decrement the loop counter , and call itself until the loop counter value is equal to zero ,

press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

NEW SEQ/MODIFY SEQ

SEQUENCE 2 SEQ2

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MEAS

TransFWD S21 (B/R)

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SCALE REF

AUTO SCALE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER F CTN

MKR SEARCH

SEARCH: MAX

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

SPECIAL FUNCTIONS

DECISION MAKING

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DECR LOOP COUNTER

IF LOOP COUNTER<> 0

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQUENCE 2 SEQ2

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

DONE SEQ MODIFY

This will create a displayed list as shown:

SEQUENCE LOOP 2

TransFWD S21 (B/R)

SCALE/DIV

AUTO SCALE

SEARCH MAX

DECR LOOP COUNTER

IF LOOP COUNTER <> 0 THEN DO

SEQUENCE 2

T o run the loop sequence , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRESET

SEQUENCE 1 SEQ1

Show 24 more labels / short text on this page
· · 10 · · · · Start of Sequence · · · · · · · · · · · · · Start of Sequence · MKR Fctn · · · 2-62 Making Measurements
Page 112 of the Agilent Technologies 8753D user manual — LOOP COUNTER
Page 112 · click to enlarge
Page 112

Generating Files in a Loop Counter Example Sequence

This example shows how to increment the names of tiles that are generated by a sequence with

a loop structure .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

NEW SEQ/MODIFY SEQ

SEQUENCE 1 SEQ 1

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SPECIAL FUNCTIONS

DECISION MAKING

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOOP COUNTER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SA VE/RECALL

SELECT DISK

INTERNAL DISK

NNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RETURN

DEFINE DISK-SAVE

DATA ONLY ON

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNN

LOCAL

SET ADDRESSES

PLOTTER PORT

DISK

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

DO SEQUENCE

SEQUENCE 2

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DONE SEQ MODIFY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NEW SEQ/MODIFY SEQ

SEQUENCE 2 SEQ 2

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FILE UTILITIES

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQUENCE FILE NAMING

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FILE NAME FILE0

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ERASE TITLE

NNNNN

D

NNNNN

T

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOOP COUNTER

NNNNNNNNNNNNNN

DONE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOT NAME PLOTFILE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ERASE TITLE

NNNNN

NNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNN

P

LOOP COUNTER

DONE

RETURN

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNN

MENU

TRIGGER MENU

SINGLE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SA VE/RECALL

SAVE STATE

NNNNNNNNNNNNNN

COPY

PLOT

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

SPECIAL FUNCTIONS

DECISION MAKING

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DECR LOOP COUNTER

IF LOOP COUNTER < > 0

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQUENCE 2 SEQ 2

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

DONE SEQ MODIFY

This will create the following displayed lists:

LOOP COUNTER

INTERNAL DISK

DATA ONLY

ON

DO SEQUENCE

SEQUENCE 2

Show 25 more labels / short text on this page
· · 7 · · · · · · · · · · Save/Recall · · · · · · · · · · · Start of Sequence · Making Measurements 2-63
Page 113 of the Agilent Technologies 8753D user manual — N
Page 113 · click to enlarge
Page 113

FILE NAME

DT[LOOP]

PLOT NAME

PL[LOOP]

SINGLE

SAVE FILE 0

PLOT

DECR LOOP COUNTER

IF LOOP COUNTER 0 THEN DO

SEQUENCE 2

Sequence 1 initializes the loop counter and calls sequence 2. Sequence 2 repeats until the loop

counter reaches 0. It takes a single sweep , saves the data le and plots the display .

The data le names generated by this sequence will be:

DT00007.D1

DT000001.D1

The plot le names generated by this sequence will be:

PL00007.FP

PL00001.FP

T o run the sequence , press:

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRESET

SEQUENCE 1 SEQ 1

Limit T est Example Sequence

This measurement example shows you how to create a sequence that will branch the sequence

according to the outcome of a limit test. Refer to \T esting a Device with Limit Lines ," located

earlier in this Chapter , for a procedure that shows you how to create a limit test.

F or this example , you must have already saved the following in register 1:

a series of active (visible) limit lines

an active limit test

1. T o create a sequence that will recall the desired instrument state , perform a limit test, and

branch to another sequence position based on the outcome of that limit test, press:

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

NEW SEQ MODIFY

SEQUENCE 1 SEQ1

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SA VE/RECALL

RECALL KEYS MENU

RECALL REG1

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

SPECIAL FUNCTIONS

DECISION MAKING

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

IF LIMIT TEST PASS

SEQUENCE 2 SEQ2

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

IF LIMIT TEST FAIL

SEQUENCE 3 SEQ3

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

DONE SEQ MODIFY

This will create a displayed list for sequence 1, as shown:

Show 16 more labels / short text on this page
Start of Sequence · through · through · · · device measurement parameters · · · · · · · · · Start of Sequence · 2-64 Making Measurements
Page 114 of the Agilent Technologies 8753D user manual — TITLE
Page 114 · click to enlarge
Page 114

RECALL REG 1

IF LIMIT TEST PASS THEN DO

SEQUENCE 2

IF LIMIT TEST FAIL THEN DO

SEQUENCE 3

2. T o create a sequence that stores the measurement data for a device that has passed the limit

test, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NEW SEQ MODIFY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQUENCE 2 SEQ2

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SELECT DISK

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INTERNAL DISK

NNNNNNNNNNNNNNNNNNNN

RETURN

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DEFINE DISK-SAVE

DATA ARRAY ON

RETURN

SAVE STATE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

DONE SEQ MODIFY

This will create a displayed list for sequence 2, as shown:

INTERNAL DISK

DATA ARRAY

ON

FILENAME

FILE 0

SAVE FILE

3. T o create a sequence that prompts you to tune a device that has failed the limit test, and

calls sequence 1 to retest the device , press:

N

NEW SEQ/MODIFY SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQUENCE 3 SEQ3

SEQ

NNNNNNNNNNNNNN

MORE

NNNNNNNNNNNNNNNNN

TITLE

DISPLA Y

NNNNN

T

NNNNN

U

NNNNN

N

NNNNN

E

NNNNN

D

NNNNN

E

NNNNN

NNNNN

NNNNN

NNNNN

E

NNNNNNNNNNNNNN

DONE

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SPECIAL FUNCTIONS

NNNNNNNNNNNNNNNNN

PAUSE

NNNNNNNNNNNNNNNNNNNN

RETURN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DO SEQUENCE

SEQUENCE 1 SEQ1

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

DONE SEQ MODIFY

This will create a displayed list for sequence 3, as shown:

TITLE

TUNE DEVICE

SEQUENCE

PAUSE

DO SEQUENCE

SEQUENCE 1

Show 17 more labels / short text on this page
· · · · · · Start of Sequence · · · · · · · · · Start of Sequence · Making Measurements 2-65
Page 115 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNN
Page 115 · click to enlarge
Page 115

The analyzer has the unique capability of measuring swept second and third harmonics as a

function of frequency in a real-time manner . Figure 2-47 displays the absolute power of the

fundamental and second harmonic in dBm. Figure 2-48 shows the second harmonic's power

level relative to the fundamental power in dBc . F ollow the steps listed below to perform these

measurements .

1. Press

CHAN 1

MEAS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TransFWD S21 (B/R)

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INPUT PORTS

NNNNN

B to measure the power for

the fundamental frequencies .

CHAN 2

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNN

MEAS

INPUT PORTS

B to measure the power for the harmonic frequencies .

3. Set the start frequency to a value greater than 16 MHz.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

COUPLED CH OFF . Uncoupling the channels allows you to have the

MENU

separate sweeps necessary for measuring the fundamental and harmonic frequencies .

NNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CHAN POWER [COUPLED] . Coupling the channel power allows you

to maintain the same fundamental frequency power level for both channels .

NNNNNNNNNNNNNNNNN

MENU

POWER and set the power level for both channels .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DISPLA Y

DUAL CHAN ON .

8. Press

and position marker to desired frequency .

MARKER

Agilent Technologies 8753D — MARKER diagram (page 115)
Fig. 53

Figure 2-47. Fundamental and 2nd Harmonic P ower Levels in dBm

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

HARMONIC MEAS

NNNNNNNNNNNNNNNNNNNN

SECOND . Y ou can view both the fundamental power and

SYSTEM

harmonic power levels at the same time . (Refer to Figure 2-47.)

NNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CHAN 2

DISPLA Y

D2/D1 toD2 ON . This display mode lets you see

the relationship between the fundamental and second or third harmonic in dBc . (Refer to

Figure 2-48.)

Show 18 more labels / short text on this page
Measuring Swept Harmonics · · Press · · Press · and select · Press · POWER and select · Press · · Press · and select · Press · · Press · · MORE and select · 2-66 Making Measurements
Page 116 of the Agilent Technologies 8753D user manual — Chapter 2 Making Measurements
Page 116 · click to enlarge
Page 116
Agilent Technologies 8753D — Chapter 2 Making Measurements diagram (page 116)
Fig. 54 · Chapter 2 Making Measurements

Figure 2-48. 2nd Harmonic P ower Level in dBc

Making Measurements 2-67

Page 117 of the Agilent Technologies 8753D user manual — Chapter 2 Making Measurements
Page 117 · click to enlarge
Page 117

Measuring a Device in the Time Domain (Option 010 Only)

The HP 8753D Option 010 allows you to measure the time domain response of a device . Time

domain analysis is useful for isolating a device problem in time or in distance . Time and

distance are related by the velocity factor of your device under test. The analyzer measures

the frequency response of your device and uses an inverse F ourier transform to convert the

data to the time domain.

Time domain analysis allows you to mathematically remove individual parts of the time domain

response to see the e ect of potential design changes . Y ou can accomplish this by \gating" out

the undesirable responses .

This section shows you how to use the time domain function to measure a device response by

the following measurement examples:

transmission measurement of RF crosstalk and multi-path signal through a surface acoustic

wave (SA W) lter

re ection measurement that locates re ections along a terminated transmission line

Transmission Response in Time Domain

In this example measurement there are three components of the transmission response:

RF leakage at near zero time

the main travel path through the device (1.6 s travel time)

the \triple travel" path (4.8 s travel time)

This example procedure also shows you how time domain analysis allows you to mathematically

remove individual parts of the time domain response to see the e ect of potential design

changes . This is accomplished by \gating" out the undesirable responses . With the \gating"

capability , the analyzer time domain allows you perform \what if" analysis by mathematically

removing selected re ections and seeing the e ect in the frequency domain.

1. Connect the device as shown in Figure 2-49.
Agilent Technologies 8753D — Chapter 2 Making Measurements diagram (page 117)
Fig. 55 · Chapter 2 Making Measurements

Figure 2-49. Device Connections for Time Domain Transmission Example Measurement

Gating · 2-68 Making Measurements

Page 118 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 118 · click to enlarge
Page 118
2. T o choose the measurement parameters , press:

PRESET

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MEAS

Trans:FWD S21 (B/R)

ST ART

M/

STOP

M/

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AUTO SCALE

SCALE REF

3. Substitute a thru for the device under test and perform a frequency response correction.

Refer to \Calibrating the Analyzer ," located at the beginning of this Chapter , for a detailed

procedure .

4. Reconnect your device under test.
5. T o transform the data from the frequency domain to the time domain and set the sweep

from 0 s to 6 s , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SYSTEM

TRANSFORM MENU

BANDPASS

TRANSFORM ON

ST ART

STOP

M/

The other time domain modes , low pass step and low pass impulse , are described in

Chapter 6, \Application and Operation Concepts ."

6. T o better view the measurement trace , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REFERENCE VALUE and turn the front panel knob or enter a value from the

SCALE REF

front panel keypad.

7. T o measure the peak response from the main path, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER F CTN

MKR SEARCH

SEARCH: MAX

The three responses shown in Figure 2-50 are the RF leakage near zero seconds , the

main travel path through the lter , and the triple travel path through the lter . Only the

combination of these responses was evident to you in the frequency domain.

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 118)
Fig. 56

Figure 2-50. Time Domain Transmission Example Measurement

Show 30 more labels / short text on this page
· · · · · · 119 · · · · 149 · · · · · · · · 0 · G/n · · · · 6 · · · · · · Making Measurements 2-69
Page 119 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 119 · click to enlarge
Page 119
8. T o access the gate function menu, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRANSFORM MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SPECIFY GATE

NNNNNNNNNNNNNNNNNNNN

CENTER

SYSTEM

9. T o set the gate parameters , by entering the marker value , press:

1.6

M/

, or turn the front panel knob to position the \>" center gate marker .

10. T o set the gate span, press:

NNNNNNNNNNNNNN

SPAN

1.2

M/

or turn the front panel knob to position the \ ag" gate markers .

11. T o activate the gating function to remove any unwanted responses , press:

NNNNNNNNNNNNNNNNNNNNNNN

GATE ON

As shown in Figure 2-51, only response from the main path is displayed.

Y ou may remove the displayed response from inside the gate markers by

N

NNNNNNNNNNNNN

SPAN and turning the front panel knob to exchange the \ ag" marker

positions .

Agilent Technologies 8753D — N diagram (page 119)
Fig. 57

Figure 2-51.

Gating in a Time Domain Transmission Example Measurement

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

T o adjust the gate shape for the best possible time domain response , press

select between minimum, normal, wide , and maximum. Each gate has a di erent passband

atness , cuto rate , and sidelobe levels .

Show 6 more labels / short text on this page
· · · pressing · GATE SHAPE and · 2-70 Making Measurements
Page 120 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 120 · click to enlarge
Page 120

T able 2-2. Gate Characteristics

Gate Span Minimum 0.1 dB 48 dB 1.4/Freq Span 2.8/Freq Span

Normal 0.1 dB 68 dB 2.8/Freq Span 5.6/Freq Span

Wide 0.1 dB 57 dB 4.4/Freq Span 8.8/Freq Span

Maximum 0.01 dB 70 dB 12.7/Freq Span 25.4/Freq Span

The passband ripple and sidelobe levels are descriptive of the gate shape . The cuto time is the

time between the stop time (6 dB on the lter skirt) and the peak of the rst sidelobe , and

is equal on the left and right side skirts of the lter . Because the minimum gate span has no

passband, it is just twice the cuto time .

Agilent Technologies 8753D — Chapter 2 Making Measurements diagram (page 120)
Fig. 58 · Chapter 2 Making Measurements

Figure 2-52. Gate Shape

13. T o see the e ect of the gating in the frequency domain, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRANSFORM MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRANSFORM OFF

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SCALE REF

AUTO SCALE

NNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DISPLA Y

DATA ! MEM

DISPLAY: DATA AND MEMORY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNN

SYSTEM

TRANSFORM MENU

SPECIFY GATE

GATE OFF

This places the gated response in memory . Figure 2-53 shows the e ect of removing the RF

leakage and the triple travel signal path using gating. By transforming back to the frequency

domain we see that this design change would yield better out-of-band rejection.

Show 19 more labels / short text on this page
Gate · P assband · Sidelobe · Cuto · Minimum · Shape · Ripple · Levels · Time · Gate Span · · · · · · · · · Making Measurements 2-71
Page 121 of the Agilent Technologies 8753D user manual — Chapter 2 Making Measurements
Page 121 · click to enlarge
Page 121
Agilent Technologies 8753D — Chapter 2 Making Measurements diagram (page 121)
Fig. 59 · Chapter 2 Making Measurements

Figure 2-53.

Gating E ects in a Frequency Domain Example Measurement

2-72 Making Measurements

Page 122 of the Agilent Technologies 8753D user manual — ST ART
Page 122 · click to enlarge
Page 122

Re ection Response in Time Domain

The time domain response of a re ection measurement is often compared with the time

domain re ectometry (TDR) measurements . Like the TDR, the analyzer measures the size of

the re ections versus time (or distance). Unlike the TDR, the time domain capability of the

analyzer allows you to choose the frequency range over which you would like to make the

measurement.

1. T o choose the measurement parameters , press:

PRESET

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MEAS

ReflFWD S11 (A/R)

ST ART

M/

STOP

2. P erform an S

1-port correction on PORT 1. Refer to Chapter 5, \Optimizing Measurement

Results ," for a detailed procedure .

3. Connect your device under test as shown in Figure 2-54.
Agilent Technologies 8753D — ST ART diagram (page 122)
Fig. 60

Figure 2-54. Device Connections for Re ection Time Domain Example Measurement

Show 14 more labels / short text on this page
· · · · · · 50 · · · · 3 · · G/n · Making Measurements 2-73
Page 123 of the Agilent Technologies 8753D user manual — F ORMA T
Page 123 · click to enlarge
Page 123
4. T o better view the measurement trace , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AUTO SCALE

SCALE REF

Figure 2-55 shows the frequency domain re ection response of the cables under test. The

complex ripple pattern is caused by re ections from the adapters interacting with each

other . By transforming this data to the time domain, you can determine the magnitude of

the re ections versus distance along the cable .

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 123)
Fig. 61

Figure 2-55. Device Response in the Frequency Domain

5. T o transform the data from the frequency domain to the time domain, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SYSTEM

TRANSFORM MENU

BANDPASS

TRANSFORM ON

6. T o view the time domain over the length ( < 4 meters) of the cable under test, press:

F ORMA T

NNNNNNNNNNNNNNNNNNNNNNN

LIN MAG

ST ART

STOP

The stop time corresponds to the length of the cable under test. The energy travels about

1 foot per nanosecond, or 0.3 meter/ns , in free space . Most cables have a relative velocity of

about 0.66 the speed in free space . Calculate about 3 ns/foot, or 10 ns/meter , for the stop

time when you are measuring the return trip distance to the cable end.

Show 16 more labels / short text on this page
· · · · · · · · 0 · · · · 35 · · G/n · 2-74 Making Measurements
Page 124 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNN
Page 124 · click to enlarge
Page 124
7. T o enter the relative velocity of the cable under test, press:

NNNNNNNNNNNNNN

MORE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

VELOCITY FACTOR

CAL

and enter a velocity factor for your cable under test.

Most cables have a relative velocity of 0.66 (for polyethylene dielectrics) or 0.7

(for te on dielectrics). If you would like the markers to read actual one-way

distance rather than return trip distance , enter one-half the actual velocity

factor . Then the markers will read the actual round trip distance to the

re ection of interest rather than the \electrical length" that assumes a relative

velocity of 1.

V el ocity F actor =

is the relative permittivity of the cable dielectric .

8. T o position the marker on the re ection of interest, press:

MARKER

and turn the front panel knob or enter a value from the front panel keypad.

In this example , the velocity factor was set to one-half the actual value , so the marker reads

the time and distance to the re ection.

9. T o position a marker at each re ection of interest, as shown in Figure 2-56 , press:

NNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER 2

NNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER 3

NNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER 4 , turning the front panel knob or entering a value from the

front panel keypad after each key press .

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 124)
Fig. 62

Figure 2-56. Device Response in the Time Domain

Show 9 more labels / short text on this page
· · p · " · r · where · r · · Making Measurements 2-75
Page 125 of the Agilent Technologies 8753D user manual — Chapter 2 Making Measurements
Page 125 · click to enlarge
Page 125

The capability of making non-coaxial measurements is available with the HP 8753 family of

analyzers with TRL* (thru-re ect-line) or LRM* (line-re ect-match) calibration. F or indepth

information on TRL*/LRM* calibration, refer to Chapter 6, \Application and Operation

Concepts ."

Non-coaxial, on-wafer measurements present a unique set of challenges for error correction in

the analyzer:

The close spacing between the microwave probes makes it dicult to maintain a high degree

of isolation between the input and the output.

The type of device measured on-wafer is often not always a simple two-port.

It may be dicult to make repeatable on-wafer contacts due to the size of the device contact

pads .

Due to the simplicity of the calibration standards , TRL* or LRM* calibrations may be used

for non-coaxial applications such as on-wafer measurements . This type of calibration with

time domain gating and a variety of probe styles can provide optimal accuracy in on-wafer

measurements . At frequencies where on-wafer calibration standards are available , SOL T

calibrations can also be done and may be preferred due to the better accuracy of the SOL T

calibration method.

F or information on how to perform TRL* or LRM* calibrations , refer to the section \TRL* and

TRM* Error-Correction" in Chapter 5, \Optimizing Measurement Results ."

Non-coaxial Measurements · 2-76 Making Measurements

Page 126 of the Agilent Technologies 8753D user manual — Chapter 3 Making Mixer Measurements
Page 126 · click to enlarge
Page 126

This chapter contains information and example procedures on the following topics:

Minimizing Source and Load Mismatches

Reducing the E ect of Spurious Responses

Eliminating Unwanted Mixing and Leakage Signals

How RF and IF Are De ned

Frequency O set Mode Operation

Di erences Between Internal and External R-Channel Inputs

P ower Meter Calibration

Conversion loss using the frequency o set mode

High dynamic range swept RF/IF conversion loss

Conversion compression using the frequency o set mode

LO to RF isolation

Where to Look for More Information

A dditional information about many of the topics discussed in this chapter is located in the

following areas:

Chapter 2, \Making Measurements ," contains step-by-step procedures for making

measurements or using particular functions .

Chapter 4, \Printing, Plotting, and Saving Measurement Results ," contains instructions

for saving to disk or the analyzer internal memory , and printing and plotting displayed

measurements .

Chapter 5, \Optimizing Measurement Results ," describes techniques and functions for

achieving the best measurement results .

Chapter 6, \Application and Operation Concepts ," contains explanatory-style information

about many applications and analyzer operation.

Making Mixer Measurements 3-1

Show 6 more labels / short text on this page
Making Mixer Measurements · Measurement considerations · Fixed IF measurements · Group delay measurements · Isolation · RF feedthrough
Page 127 of the Agilent Technologies 8753D user manual — Chapter 3 Making Mixer Measurements
Page 127 · click to enlarge
Page 127

T o ensure successful mixer measurements , the following measurement challenges must be taken

into consideration:

Minimizing Source and Load Mismatches

Reducing the E ect of Spurious Responses

Eliminating Unwanted Mixing and Leakage Signals

How RF and IF Are De ned

Frequency O set Mode Operation

Di erences Between Internal and External R-Channel Inputs

P ower Meter Calibration

Minimizing Source and Load Mismatches

When characterizing linear devices , you can use vector accuracy enhancement to

mathematically remove all systematic errors , including source and load mismatches , from your

measurement. This is not possible when the device you are characterizing is a mixer operating

over multiple frequency ranges . Therefore , source and load mismatches are not corrected for

and will add to overall measurement uncertainty .

Y ou should place attenuators at all of the test ports to reduce the measurement errors

associated with the interaction between mixer port matches and system port matches . T o avoid

overdriving the receiver , you should give extra care to selecting the attenuator located at the

mixer's IF port. F or best results , you should choose the attenuator value so that the power

incident on the analyzer R-channel input is less than 10 dBm and greater than 35 dBm.

Reducing the E ect of Spurious Responses

By choosing test frequencies (frequency list mode), you can reduce the e ect of spurious

responses on measurements by avoiding frequencies that produce IF signal path distortion.

Eliminating Unwanted Mixing and Leakage Signals

By placing lters between the mixer's IF port and the receiver's input port, you can eliminate

unwanted mixing and leakage signals from entering the analyzer's receiver . Filtering is

required in both xed and broadband measurements . Therefore , when con guring broad-band

(swept) measurements you may need to trade some measurement bandwidth for the ability to

more selectively lter signals entering the analyzer receiver .

How RF and IF Are De ned

In standard mixer measurements , the input of the mixer is always connected to the analyzer's

RF source , and the output of the mixer always produces the IF frequencies that are received by

the analyzer's receiver .

However , the ports labeled RF and IF on most mixers are not consistently connected to the

analyzer's source and receiver ports , respectively . These mixer ports are switched, depending

on whether a down converter or an up converter measurement is being performed.

It is important to keep in mind that in the setup diagrams of the frequency o set mode , the

analyzer's source and receiver ports are labeled according to the mixer port that they are

connected to .

3-2 Making Mixer Measurements

Measurement Considerations · Mixer Considerations · Analyzer Operation

Page 128 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNN
Page 128 · click to enlarge
Page 128

In a down converter measurement where the

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DOWN CONVERTER softkey is selected, the

labeled RF , connecting to the mixer RF port, and the analyzer's receiver frequency is labeled

IF , connecting to the mixer IF port.

Because the RF frequency can be greater or less than the set LO frequency in this type of

NNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNN

measurement, you can select either

RF > LO or

RF < LO .

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNNNNN diagram (page 128)
Fig. 63

Figure 3-1. Down Converter P ort Connections

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

In an up converter measurement where the

UP CONVERTER softkey is selected, the notation

on the setup diagram indicates that the analyzer's source frequency is labeled IF , connecting

to the mixer IF port, and the analyzer's receiver frequency is labeled RF , connecting to the

mixer RF port.

Because the RF frequency will always be greater than the set LO frequency in this type of

measurement, you must select only

NNNNNNNNNNNNNNNNNNNNNNN

RF > LO .

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNNNNN diagram (page 128)
Fig. 64

Figure 3-2. Up Converter P ort Connections

Making Mixer Measurements 3-3

Page 129 of the Agilent Technologies 8753D user manual — MENU
Page 129 · click to enlarge
Page 129

Frequency O set Mode Operation

Frequency o set measurements do not begin until all of the frequency o set mode parameters

are set. These include the following:

Start and Stop IF Frequencies

Up Converter / Down Converter

RF > LO / RF < LO

The LO frequency for frequency o set mode must be set to the same value as the external LO

source . The o set frequency between the analyzer source and receiver will be set to this value .

When frequency o set mode operation begins , the receiver locks onto the entered IF signal

frequencies and then o sets the source frequency required to produce the IF . Therefore , since

it is the analyzer receiver that controls the source , it is only necessary to set the start and stop

frequencies from the receiver .

Di erences Between Internal and External R-Channel Inputs

Due to internal losses in the analyzer's test set, the power measured internally at the

R-Channel is 16 dB lower than that of the source . T o compensate for these losses , the traces

associated with the R-Channel have been o set 16 dB higher . As a result, power measured

directly at the R-Channel via the R CHANNEL IN port will appear to be 16 dB higher than its

actual value . If power meter calibration is not used, this o set in power must be accounted for

with a receiver calibration before performing measurements .

The following steps can be performed to observe this o set in power:

1. T o set the power range to manual, press:

MENU

NNNNNNNNNNNNNNNNN

POWER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PWR RANGE MAN

Setting the power range to manual prevents the internal source attenuator from switching

when changing power levels . If you choose a di erent power range , the R-channel o set

compensation and R-channel measurement changes by the amount of the attenuator setting.

2. Connect the analyzer source output, port 1, directly to the R-channel input as shown in

Figure 3-3.

T o prevent connector damage , use an adapter (HP part number 1250-1462) as a

connector saver for R CHANNEL IN.

Agilent Technologies 8753D — MENU diagram (page 129)
Fig. 65

Figure 3-3. R-Channel External Connection

3-4 Making Mixer Measurements

Show 5 more labels / short text on this page
LO frequency · · · 0 ·
Page 130 of the Agilent Technologies 8753D user manual — CAL
Page 130 · click to enlarge
Page 130
3. T o activate the frequency o set mode , press:

SYSTEM

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INSTRUMENT MODE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FREQ OFFS MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FREQ OFFS ON

Since the LO (o set) frequency is still set to the default value of 0 Hz, the analyzer will

operate normally .

4. Measure the output power in the R-channel by pressing:

MEAS

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INPUT PORTS

NNNNN

R

Observe the 13 to 16 dB o set in measured power . The actual input power level to the

R-channel input must be 0 dBm or less , 10 dBm typical, to avoid receiver saturation

e ects . The minimum signal level must be greater than 35 dBm to provide sucient signal

for operation of the phaselock loop .

5. Y ou cannot trust R channel power settings without knowing about the o set involved.

P erform a receiver calibration to remove any power o sets by pressing:

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECEIVER CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TAKE RCVR CAL SWEEP

Once completed, the R-channel should display 0 dBm. Changing power ranges will require a

recalibration of the R-channel.

Making Mixer Measurements 3-5

Show 6 more labels / short text on this page
· · · · ·
Page 131 of the Agilent Technologies 8753D user manual — OR
Page 131 · click to enlarge
Page 131

P ower Meter Calibration

Mixer transmission measurements are generally con gured as follows:

measured output power (W atts) / set input power (W atts)

OR

measured output power (dBm) set input power (dBm)

F or this reason, the set input power must be accurately controlled in order to ensure

measurement accuracy .

The amplitude variation of the analyzer is speci ed 1 dB over any given source frequency .

This may give a maximum 2 dB error for a mixer transmission test setup: 1 dB for the source

over the IF range during measurement and 1 dB over the RF range during measurement.

Higher measurement accuracy may be obtained through the use of power meter calibration.

Y ou can use power meter calibration to correct for power o sets , losses , and atness variations

occurring between the analyzer source and the input to the mixer under test.

3-6 Making Mixer Measurements

Page 132 of the Agilent Technologies 8753D user manual — MENU
Page 132 · click to enlarge
Page 132

Conversion Loss Using the Frequency O set Mode

Conversion loss is the measure of eciency of a mixer . It is the ratio of side-band IF power

to RF signal power , and is usually expressed in dB . (Express ratio values in dB amounts to a

subtraction of the dB power in the denominator from the dB power in the numerator .) The

mixer translates the incoming signal, (RF), to a replica, (IF), displaced in frequency by the

local oscillator , (LO). Frequency translation is characterized by a loss in signal amplitude and

the generation of additional sidebands . F or a given translation, two equal output signals are

expected, a lower sideband and an upper sideband.

Agilent Technologies 8753D — Chapter 3 Making Mixer Measurements diagram (page 132)
Fig. 66 · Chapter 3 Making Mixer Measurements

Figure 3-4.

An Example Spectrum of RF , LO , and IF Signals Present in a

The analyzer allows you to make a swept RF/IF conversion loss measurement holding the LO

frequency xed. Y ou can make this measurement by using the analyzer's frequency o set

measurement mode . This mode of operation allows you to o set the analyzer's source by a

xed value , above or below the analyzer's receiver . That is , this allows you to use a device

input frequency range that is di erent from the receiver input frequency range .

The following procedure describes the swept IF frequency conversion loss measurement of a

broadband component mixer:

1. Set the LO source to the desired CW frequency and power level.

CW frequency = 1000 MHz

P ower = 13 dBm

2. Set the desired source power to the value which will provide 10 dBm or less to the

R-channel input. Press:

MENU

NNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PWR RANGE MAN

POWER

3. Calibrate and zero the power meter .
4. Connect the measurement equipment as shown in Figure 3-5.

The low pass lter is required to limit the range of frequencies passed into the R-channel

input port. The lter is selected to pass the IF frequencies for the measurement but

prevent the LO feedthrough and unwanted mixer products from confusing the phase lock

loop operation.

A pad is used to isolate the lter and improve the IF port match for the mixer .

The attenuation of the power splitter is used to improve the RF port match for the mixer .

Making Mixer Measurements 3-7

Show 5 more labels / short text on this page
Conversion Loss Measurement · · · 0 ·
Page 133 of the Agilent Technologies 8753D user manual — NNNNNNNNNNN
Page 133 · click to enlarge
Page 133

T o prevent connector damage , use an adapter (HP part number 1250-1462) as a

connector saver for R CHANNEL IN.

Agilent Technologies 8753D — Chapter 3 Making Mixer Measurements diagram (page 133)
Fig. 67 · Chapter 3 Making Mixer Measurements

Figure 3-5. Connections for R Channel and Source Calibration

5. From the front panel of the HP 8753D or analyzer , set the desired receiver frequency and

source output power , by pressing:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INSTRUMENT MODE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FREQ OFFS MENU

SYSTEM

ST ART

M/

STOP

M/

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FREQ OFFS ON

MENU

NNNNNNNNNNNNNNNNN

POWER

6. T o view the measurement trace , press:

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNN

MEAS

INPUT PORTS

R

7. Select the HP 8753D as the system controller:

LOCAL

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SYSTEM CONTROLLER

8. Set the power meter's address:

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SET ADDRESSES

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ADDRESS: P MTR/HPIB

Select the appropriate power meter by pressing

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

POWER MTR [ ] until the correct model

number is displayed (HP 436A or HP 438A/437).

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL

PWRMTR CAL

LOSS/SENSR LISTS

CAL FACTOR SENSOR A and enter the

correction factors as listed on the power sensor . Press

NNNNNNNNNNN

ADD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FREQUENCY

M/

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNN

NNNNNNNNNNNNNN

CAL FACTOR

DONE for each correction factor . When nished, press

DONE .

3-8 Making Mixer Measurements

Show 27 more labels / short text on this page
· · · · 100 · · · · 350 · · · · 0 · · · · · · · · ## · Press · · · · ·
Page 134 of the Agilent Technologies 8753D user manual — SYSTEM
Page 134 · click to enlarge
Page 134
11. T o perform a one sweep power meter calibration over the IF frequency range at 0 dBm,

press:

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PWRMTR CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ONE SWEEP

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TAKE CAL SWEEP

Because power meter calibration requires a longer sweep time , you may want

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

to reduce the number of points before pressing

TAKE CAL SWEEP . After the

power meter calibration is nished, return the number of points to its original

value and the analyzer will automatically interpolate this calibration.

12. T o calibrate the R-channel over the IF range , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL

RECEIVER CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TAKE RCVR CAL SWEEP

Once completed, the display should read 0 dBm.

13. Make the connections as shown in Figure 3-6 for the one-sweep power meter calibration

over the RF range .

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 134)
Fig. 68

Figure 3-6.

Connections for a One-Sweep P ower Meter Calibration for Mixer Measurements

14. T o set the frequency o set mode LO frequency from the analyzer , press:

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INSTRUMENT MODE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FREQ OFFS MENU

NNNNNNNNNNNNNNNNNNNNNNN

LO MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FREQUENCY:CW

M/

Making Mixer Measurements 3-9

Show 11 more labels / short text on this page
· · · 0 · · · · · · 1000 ·
Page 135 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 135 · click to enlarge
Page 135
15. T o select the converter type and a high-side LO measurement con guration, press:

NNNNNNNNNNNNNNNNNNNN

RETURN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DOWN CONVERTER

NNNNNNNNNNNNNNNNN

RF<LO

actually sweeping backwards , as shown in Figure 3-7. The measurements setup diagram is

shown in Figure 3-8 .

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNN diagram (page 135)
Fig. 69

Figure 3-7. Diagram of Measurement Frequencies

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNN diagram (page 135)
Fig. 70

Figure 3-8. Measurement Setup from Display

16. T o view the measurement trace , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

VIEW MEASURE

17. T o perform a one-sweep power meter calibration over the RF frequency range , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ONE SWEEP

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL

PWRMTR CAL

TAKE CAL SWEEP

Do not reduce the number of points to perform this power meter calibration.

Reducing the number of points will turn o the receiver calibration.

The analyzer is now displaying the conversion loss of the mixer calibrated with power

meter accuracy .

3-10 Making Mixer Measurements

Show 4 more labels / short text on this page
· · 0 ·
Page 136 of the Agilent Technologies 8753D user manual — N
Page 136 · click to enlarge
Page 136
18. T o view the conversion loss in the best vertical resolution, press:

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNN

SCALE REF

AUTOSCALE

Agilent Technologies 8753D — N diagram (page 136)
Fig. 71

Figure 3-9. Conversion Loss Example Measurement

C onv er sion l oss=g ain = output pow er input pow er

In this measurement, you set the input power and measured the output power . Figure 3-9

shows the absolute loss through the mixer versus mixer output frequency . If the mixer

under test contained built-in ampli cation, then the measurement results would have shown

conversion gain.

Making Mixer Measurements 3-11

·

Page 137 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNN
Page 137 · click to enlarge
Page 137

High Dynamic Range Swept RF/IF Conversion Loss

The HP 8753D's frequency o set mode enables the testing of high dynamic range frequency

converters (mixers), by tuning the analyzer's high dynamic range receiver above or below its

source , by a xed o set. This capability allows the complete measurement of both pass and

reject band mixer characteristics .

The analyzer has a 35 dB dynamic range limitation on measurements made directly with its R

(phaselock) channel. F or this reason, the measurement of high dynamic range mixing devices

(such as mixers with built in ampli cation and ltering) with greater than 35 dB dynamic range

must be made on either the analyzer's A or B channel, with a reference mixer providing input

to the analyzer's R-channel for phaselock.

This example describes the swept IF conversion loss measurement of a mixer and lter . The

output ltering demonstrates the analyzer's ability to make high dynamic range measurements .

T o avoid the complexity of performing a separate power meter calibration over the RF

frequency range while the mixer under test and reference mixer are operating, a broad band

power meter calibration is used. The broad band calibration covers the entire range of IF and

RF frequencies .

1. Set the following analyzer parameters:

PRESET

ST ART

M/

STOP

2.5

NNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PWR RANGE MAN

MENU

POWER

2. Calibrate and zero the power meter .
3. Connect the measurement equipment as shown in Figure 3-10.

T o prevent connector damage , use an adapter (HP part number 1250-1462) as a

connector saver for R CHANNEL IN.

3-12 Making Mixer Measurements

Show 14 more labels / short text on this page
· · · · 100 · · · · G/n · · · · 0 ·
Page 138 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 138 · click to enlarge
Page 138
Agilent Technologies 8753D — Chapter 3 Making Mixer Measurements diagram (page 138)
Fig. 72 · Chapter 3 Making Mixer Measurements

Figure 3-10. Connections for Broad Band P ower Meter Calibration

4. Select the HP 8753D as the system controller:

LOCAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SYSTEM CONTROLLER

5. Set the power meter's address:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SET ADDRESSES

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ADDRESS: P MTR/HPIB

Select the appropriate power meter by pressing

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

POWER MTR [ ] until the correct model

number is displayed (HP 436A or HP 438A/437).

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL

PWRMTR CAL

LOSS/SENSR LISTS

CAL FACTOR SENSOR A and enter the

correction factors as listed on the power sensor . Press

NNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FREQUENCY

ADD

M/

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL FACTOR

NNNNNNNNNNNNNN

DONE for each correction factor . When nished, press

NNNNNNNNNNNNNN

DONE

NNNNNNNNNNNNNNNNNNNN

RETURN .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNN

MEAS

INPUT PORTS

B

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL

PWRMTR CAL

10. P erform a one sweep power meter calibration over the IF frequency range at 0 dBm:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ONE SWEEP

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TAKE CAL SWEEP

Making Mixer Measurements 3-13

Show 18 more labels / short text on this page
· · · · ## · Press · · · · · · · · · · · 0 ·
Page 139 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 139 · click to enlarge
Page 139

Because power meter calibration requires a longer sweep time , you may want

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

to reduce the number of points before pressing

TAKE CAL SWEEP . After the

power meter calibration is nished, return the number of points to its original

value and the analyzer will automatically interpolate this calibration.

11. Connect the measurement equipment as shown in Figure 3-11.
Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 139)
Fig. 73

Figure 3-11. Connections for Receiver Calibration

12. Set the following analyzer parameters:

ST ART

M/

STOP

13. T o calibrate the B-channel over the IF range , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECEIVER CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TAKE RCVR CAL SWEEP

CAL

Once completed, the analyzer should display 0 dBm.

14. Make the connections shown in Figure 3-12.
15. Set the LO source to the desired CW frequency and power level. F or this example the

values are as follows:

CW frequency = 1500 MHz

source power = 13 dBm

3-14 Making Mixer Measurements

Show 11 more labels / short text on this page
· · 100 · · · · 1 · G/n · · ·
Page 140 of the Agilent Technologies 8753D user manual — Chapter 3 Making Mixer Measurements
Page 140 · click to enlarge
Page 140
Agilent Technologies 8753D — Chapter 3 Making Mixer Measurements diagram (page 140)
Fig. 74 · Chapter 3 Making Mixer Measurements

Figure 3-12.

Connections for a High Dynamic Range Swept IF Conversion Loss Measurement

Making Mixer Measurements 3-15

Page 141 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 141 · click to enlarge
Page 141
16. T o set the frequency o set mode LO frequency , press:

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INSTRUMENT MODE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FREQ OFFS MENU

NNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LO MENU

FREQUENCY:CW

M/

17. T o select the converter type and low-side LO measurement con guration, press:

NNNNNNNNNNNNNNNNNNNN

RETURN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DOWN CONVERTER

NNNNNNNNNNNNNNNNN

RF>LO

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FREQ OFFS ON

In this low-side LO , down converter measurement, the analyzer's source frequency range

will be o set higher than the receiver frequency range . The source frequency range can be

determined from the following equation:

receiver frequency range (100 { 1000 MHz) + LO frequency (1500 MHz) = 1.6-2.5 GHz

18. T o view the conversion loss in the best vertical resolution, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

VIEW MEASURE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AUTOSCALE

SCALE REF

Figure 3-13 shows the conversion loss of this low-side LO , mixer with output ltering.

dynamic range limit of the R Channel. If the mixer under test also contained ampli cation,

then this dynamic range would have been even greater due to the conversion gain of the

mixer .

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 141)
Fig. 75

Figure 3-13. Example of Swept IF Conversion Loss Measurement

3-16 Making Mixer Measurements

Show 6 more labels / short text on this page
· · 1500 · · ·
Page 142 of the Agilent Technologies 8753D user manual — Chapter 3 Making Mixer Measurements
Page 142 · click to enlarge
Page 142

Fixed IF Mixer Measurements

A xed IF can be produced by using both a swept RF and LO that are o set by a certain

frequency . With proper ltering, only this o set frequency will be present at the IF port of the

mixer .

This measurement requires two external RF sources as stimuli. Figure 3-15 shows the hardware

con guration for the xed IF conversion loss measurement. This example measurement

procedure uses the analyzer's test sequence function for automatically controlling the two

external synthesizers (with SCPI commands), and making a conversion loss measurement in

tuned receiver mode . The test sequence function is an instrument automation feature internal

to the analyzer . F or more information on the test sequence function refer to \T est Sequencing"

located in Chapter 2.

The analyzer's tuned receiver mode allows you to tune its receiver to an arbitrary frequency

and measure signal power . This is only possible if the signal you wish to analyze is at an exact

known frequency . Therefore , the RF and LO sources must be synthesized and synchronized

with the analyzer's time base .

The following sequence initializes and calibrates the network analyzer . It then initializes the

two external sources prior to measurement. This sequence includes:

putting the network analyzer into tuned receiver mode

setting up a frequency list sweep of 26 points

performing a response calibration

prompting the user to connect a mixer to the test set up

initializing a loop counter value to 26

addressing and con guring the two sources

calling the next measurement sequence

1. Make the following connections as shown in Figure 3-14 . Set the HP-IB address of the

external RF source to 19 and the external LO source to 21.

2. Con rm that the external sources are con gured to receive commands in the SCPI

programming language and that their output power is switched on.

Y ou may have to consult the User's Guide of the external source being used to

determine how to set the source to receive SCPI commands .

3. Be sure to connect the 10 MHz reference signals of the external sources to the EXT REF

connector on the rear panel of the analyzer (a BNC tee must be used).

If the 10 MHz reference signals of the external sources are connected together ,

then it will only be necessary to connect one reference signal from one of the

sources to the EXT REF connector of the analyzer .

Making Mixer Measurements 3-17

Tuned Receiver Mode · Sequence 1 Setup

Page 143 of the Agilent Technologies 8753D user manual — N
Page 143 · click to enlarge
Page 143
Agilent Technologies 8753D — Chapter 3 Making Mixer Measurements diagram (page 143)
Fig. 76 · Chapter 3 Making Mixer Measurements

Figure 3-14. Connections for a Response Calibration

4. Press the following keys on the analyzer to create sequence 1:

T o enter the following sequence commands that require titling, an external

keyboard may be used for convenience .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NEW SEQ/MODIFY SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQUENCE 1 SEQ1

SEQ

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SA VE/RECALL

SELECT DISK

INTERNAL MEMORY

NNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RETURN (Select the preset state .)

RECALL STATE

Putting the Analyzer into Tuned Receiver Mode

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOCAL

SYSTEM CONTROLLER

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SYSTEM

INSTRUMENT MODE

TUNED RECEIVER

Setting Up a Frequency List Sweep of 26 P oints

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SWEEP TYPE MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EDIT LIST

NNNNNNNNNNN

ADD

MENU

NNNNNNNNNNNNNNNNNNNNNNN

CW FREQ

M/

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NUMBER OF POINTS

NNNNNNNNNNNNNN

DONE

NNNNNNNNNNNNNN

DONE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LIST FREQ

P erforming a Response Calibration

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNN

MEAS

INPUT PORTS

B

NNNNNNNNNNNNNN

MORE

NNNNNNNNNNNNNNNNN

TITLE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ERASE TITLE

DISPLA Y

3-18 Making Mixer Measurements

Show 19 more labels / short text on this page
· · Presetting the Instrument · · · · · · · · · 100 · · 26 · · · · ·
Page 144 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNN
Page 144 · click to enlarge
Page 144

POW:LEV 6DBM

NNNNNNNNNNNNNN

DONE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

SPECIAL FUNCTIONS

PERIPHERAL HPIB ADDR

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TITLE TO PERIPHERAL

NNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DISPLA Y

MORE

TITLE

ERASE TITLE

FREQ:MODE CW;CW 100MHZ

NNNNNNNNNNNNNN

DONE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SPECIAL FUNCTIONS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PERIPHERAL HPIB ADDR

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TITLE TO PERIPHERAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CALIBRATE MENU

NNNNNNNNNNNNNNNNNNNNNNNNNN

RESPONSE

CAL

NNNNNNNNNNNNNN

THRU

Prompting the User to Connect a Mixer to the T est Set Up

N

NNNNNNNNNNNNN

NNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DISPLA Y

MORE

TITLE

ERASE TITLE

CONNECT MIXER

NNNNNNNNNNNNNN

DONE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

SPECIAL FUNCTIONS

NNNNNNNNNNNNNNNNN

PAUSE

Initializing a Loop Counter V alue to 26

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

SPECIAL FUNCTIONS

DECISION MAKING

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOOP COUNTER

SCALE REF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REFERENCE POSITION

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REFERENCE VALUE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MENU

TRIGGER MENU

MANUAL TRG ON POINT

A ddressing and Con guring the Two Sources

NNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DISPLA Y

MORE

TITLE

ERASE TITLE

FREQ:MODE CW;CW 500MHZ;:FREQ:CW:STEP 100MHZ

NNNNNNNNNNNNNN

DONE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

SPECIAL FUNCTIONS

PERIPHERAL HPIB ADDR

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TITLE TO PERIPHERAL

NNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DISPLA Y

MORE

TITLE

ERASE TITLE

POW:LEV 13DBM

NNNNNNNNNNNNNN

DONE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

SPECIAL FUNCTIONS

PERIPHERAL HPIB ADDR

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TITLE TO PERIPHERAL

NNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DISPLA Y

MORE

TITLE

ERASE TITLE

FREQ:MODE CW;CW 600MHZ;:FREQ:CW:STEP 100MHZ

NNNNNNNNNNNNNN

DONE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PERIPHERAL HPIB ADDR

SEQ

SPECIAL FUNCTIONS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TITLE TO PERIPHERAL

Making Mixer Measurements 3-19

Show 49 more labels / short text on this page
· · 19 · · · · · · 19 · · · · · · · · · · 26 · · · · 2 · · 0 · · 20 · · · · · · · · · 19 · · · · · · 21 · · · · · · 21 ·
Page 145 of the Agilent Technologies 8753D user manual — SEQUENCE SEQ1
Page 145 · click to enlarge
Page 145

Calling the Next Measurement Sequence

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

DO SEQUENCE

SEQUENCE 2 SEQ2

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

DONE SEQ MODIFY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

NEW SEQ/MODIFY SEQ

SEQUENCE 1 SEQ 1 and the analyzer will display the

following sequence commands:

SEQUENCE SEQ1

RECALL PRST STATE

SYSTEM CONTROLLER

TUNED RECEIVER

EDIT LIST

ADD

CW FREQ

NUMBER OF POINTS

DONE

DONE

LIST FREQ

B

TITLE

POW:LEV 6DBM

PERIPHERAL HPIB ADDR

TITLE TO PERIPHERAL

TITLE

FREQ:MODE CW;CW 100MHZ

TITLE TO PERIPHERAL

CALIBRATE: RESPONSE

CAL STANDARD

DONE CAL CLASS

TITLE

CONNECT MIXER

PAUSE

LOOP COUNTER

SCALE/DIV

REFERENCE POSITION

REFERENCE VALUE

MANUAL TRG ON POINT

TITLE

FREQ:MODE CW;CW 500MHZ;:FREQ:CW:STEP 100MHZ

TITLE TO PERIPHERAL

TITLE

POW:LEV 13DBM

PERIPHERAL HPIB ADDR

TITLE TO PERIPHERAL

TITLE

3-20 Making Mixer Measurements

Show 9 more labels / short text on this page
· · · · Press · · Start of Sequence · 100M/u · 20x1
Page 146 of the Agilent Technologies 8753D user manual — SEQUENCE SEQ2
Page 146 · click to enlarge
Page 146

FREQ:MODE CW;CW 600MHZ;:FREQ:CW:STEP 100MHZ

TITLE TO PERIPHERAL

DO SEQUENCE

SEQUENCE 2

The following sequence makes a series of measurements until all 26 CW measurements are

made and the loop counter value is equal to zero . This sequence includes:

incrementing the source frequencies

decrementing the loop counter

1. Press the following keys on the analyzer to create sequence 2:

T o enter the following sequence commands that require titling, an external

keyboard may be used for convenience .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

DONE SEQ MODIFY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

NEW SEQ/MODIFY SEQ

SEQUENCE 2 SEQ2

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SPECIAL FUNCTIONS

NNNNNNNNNNNNNNNNNNNN

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRIGGER MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MANUAL TRG ON POINT

MENU

Incrementing the Source Frequencies

NNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DISPLA Y

MORE

TITLE

ERASE TITLE

FREQ:CW UP

NNNNNNNNNNNNNN

DONE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SPECIAL FUNCTIONS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PERIPHERAL HPIB ADDR

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TITLE TO PERIPHERAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PERIPHERAL HPIB ADDR

TITLE TO PERIPHERAL

Decrementing the Loop Counter

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DECISION MAKING

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DECR LOOP COUNTER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

IF LOOP COUNTER<>0

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQUENCE 2 SEQ2

NNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DISPLA Y

MORE

TITLE

ERASE TITLE

MEASUREMENT COMPLETED

NNNNNNNNNNNNN

N

DONE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

DONE SEQ MODIFY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

NEW SEQ/MODIFY SEQ

SEQUENCE 2 SEQ 2 and the analyzer will display the

following sequence commands:

SEQUENCE SEQ2

Making Mixer Measurements 3-21

Show 32 more labels / short text on this page
Sequence 2 Setup · taking data · labeling the screen · · · · · T aking Data · · · WAIT x · · · · · · · · · 19 · · 21 · · Labeling the Screen · · · · · Press · · Start of Sequence · WAIT x
Page 147 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 147 · click to enlarge
Page 147

MANUAL TRG ON POINT

TITLE

FREQ:CW UP

PERIPHERAL HPIB ADDR

TITLE TO PERIPHERAL

PERIPHERAL HPIB ADDR

TITLE TO PERIPHERAL

DECR LOOP COUNTER

IF LOOP COUNTER <>0 THEN DO

SEQUENCE 2

TITLE

MEASUREMENT COMPLETED

2. Press the following keys to run the sequences:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

DONE SEQ MODIFY

DO SEQUENCE

SEQUENCE2 SEQ2

When the prompt CONNECT MIXER appears , connect the equipment as shown in Figure 3-15.

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 147)
Fig. 77

Figure 3-15. Connections for a Conversion Loss Using the Tuned Receiver Mode

When the sequences are nished you should have a result as shown in Figure 3-16.

3-22 Making Mixer Measurements

·

Page 148 of the Agilent Technologies 8753D user manual — Chapter 3 Making Mixer Measurements
Page 148 · click to enlarge
Page 148
Agilent Technologies 8753D — Chapter 3 Making Mixer Measurements diagram (page 148)
Fig. 78 · Chapter 3 Making Mixer Measurements

Figure 3-16. Example Fixed IF Mixer Measurement

The displayed trace represents the conversion loss of the mixer at 26 points . Each point

corresponds to one of the 26 di erent sets of RF and LO frequencies that were used to create

the same xed IF frequency .

Making Mixer Measurements 3-23

Page 149 of the Agilent Technologies 8753D user manual — CENTER
Page 149 · click to enlarge
Page 149

Phase or Group Delay Measurements

F or information on group delay principles , refer to \Group Delay Principles" in Chapter 6.

The accuracy of this measurement depends on the quality of the mixer that is being used for

calibration and how well this mixer has been characterized. The following measurement must

be performed with a broadband calibration mixer that has a known group delay . The following

table lists the speci cations of two mixers that may be used for calibration:

ANZA C

.03 to 3 GHz .5 ns

MCD-123

dc to 1250 MHz .6 ns

ZFM-4

1. Set the LO source to the desired CW frequency and power level. F or this example the LO

source is set to the following values:

CW frequency = 1000 MHz

power = 13 dBm

2. Initialize the analyzer by pressing

.

PRESET

3. From the front panel of the HP 8753D , set the desired receiver frequency and source

output power by pressing:

CENTER

M/

SP AN

M/

NNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MENU

POWER

PWR RANGE MAN

4. Connect the instruments as shown in Figure 3-17, placing a broadband \calibration" mixer

(ZFM-4) between PORT 1 and PORT 2.

T o prevent connector damage , use an adapter (HP part number 1250-1462) as a

connector saver for R CHANNEL IN.

3-24 Making Mixer Measurements

Show 19 more labels / short text on this page
Model · Useful · Group · Number · Frequency Range · Delay · Mini-Circuits · · · 300 · · · · 100 · · · · 0 ·
Page 150 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNN
Page 150 · click to enlarge
Page 150
Agilent Technologies 8753D — Chapter 3 Making Mixer Measurements diagram (page 150)
Fig. 79 · Chapter 3 Making Mixer Measurements

Figure 3-17. Connections for a Group Delay Measurement

5. T o set the frequency o set mode LO frequency from the analyzer , press:

SYSTEM

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INSTRUMENT MODE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FREQ OFFS MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

VIEW MEASURE

NNNNNNNNNNNNNNNNNNNNNNN

LO MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FREQUENCY:CW

M/

6. T o select the converter type and a high-side LO measurement con guration, press:

N

NNNNNNNNNNNNNNNNNNN

RETURN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DOWN CONVERTER

NNNNNNNNNNNNNNNNN

RF<LO

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FREQ OFFS ON

7. T o select the format type , press:

NNNNNNNNNNNNNNNNN

F ORMA T

DELAY

Making Mixer Measurements 3-25

Show 6 more labels / short text on this page
· · 1000 · · ·
Page 151 of the Agilent Technologies 8753D user manual — SCALE REF
Page 151 · click to enlarge
Page 151
8. T o make a response error-correction, press:

MEAS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INPUT PORTS

NNNNNNNNNNN

B/R

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNN

CAL

CALIBRATE MENU

RESPONSE

THRU

9. Replace the \calibration" mixer with the device under test. If measuring group delay , set

the delay equal to the \calibration" mixer delay (for example 0.6 ns) by pressing:

SCALE REF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ELECTRICAL DELAY

0.6

10. Scale the data for best vertical resolution.

SCALE REF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AUTOSCALE

Agilent Technologies 8753D — SCALE REF diagram (page 151)
Fig. 80

Figure 3-18. Group Delay Measurement Example

The displayed measurement trace shows the device under test delay , relative to the

\calibration" mixer . This measurement is also useful when the device under test has built-in

ltering, which requires >30 dB range (the maximum of R input). PORT 1 to PORT 2 range is

>100 dB .

3-26 Making Mixer Measurements

Show 12 more labels / short text on this page
· · · · · · · · G/n · · ·
Page 152 of the Agilent Technologies 8753D user manual — N
Page 152 · click to enlarge
Page 152

Amplitude and Phase Tracking

Using the same measurement set-up as in \Phase or Group Delay Measurements ," you can

determine how well two mixers track each other in terms of amplitude and phase .

1. Repeat steps 1 through 8 of the previous \Group Delay Measurements" section with the

following exception:

In step 7, select

NNNNNNNNNNNNNNNNN

PHASE .

F ORMA T

Once the analyzer has displayed the measurement results , press

N

DATA !MEM .

NNNNNNNNNNNNNNNNNNNNNNNNNNN

DISPLA Y

3. Replace the calibration mixer with the mixer under test.

N

NNNNNNNNNNNNNNNNNNNNNNNNN

DATA/MEM .

The resulting trace should represent the amplitude and phase tracking of the two mixers .

Making Mixer Measurements 3-27

· · Press

Page 153 of the Agilent Technologies 8753D user manual — MENU
Page 153 · click to enlarge
Page 153

Conversion Compression Using the Frequency O set Mode

Conversion compression is a measure of the maximum RF input signal level, where the mixer

provides linear operation. The conversion loss is the ratio of the IF output level to the RF input

level. This value remains constant over a speci ed input power range . When the input power

level exceeds a certain maximum, the constant ratio between IF and RF power levels will begin

to change . The point at which the ratio has decreased 1 dB is called the 1 dB compression

point. See Figure 3-19.

Agilent Technologies 8753D — Chapter 3 Making Mixer Measurements diagram (page 153)
Fig. 81 · Chapter 3 Making Mixer Measurements

Figure 3-19.

Conversion Loss and Output P ower as a Function of Input P ower Level Example

compression begins and the mixer saturates .

The following example uses a ratio of mixer output to input power and a marker search

function to locate a mixer's 1 dB compression point.

1. Set the LO source to the desired CW frequency and power level.

CW frequency = 600 MHz

P ower = 13 dBm

2. Initialize the analyzer by pressing

PRESET

.

3. T o set the desired CW frequency and power sweep range , press:

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SWEEP TYPE MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

POWER SWEEP

NNNNNNNNNNNNNNNNNNNN

RETURN

NNNNNNNNNNNNNNNNNNNNNNN

CW FREQ

M/

NNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

POWER

PWR RANGE MAN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNN

POWER RANGES

RANGE 0

ST ART

STOP

4. Make the connections , as shown in Figure 3-20.

3-28 Making Mixer Measurements

Show 14 more labels / short text on this page
· · · 800 · · · · 10 · · · · · 10 ·
Page 154 of the Agilent Technologies 8753D user manual — DISPLA Y
Page 154 · click to enlarge
Page 154

T o prevent connector damage , use an adapter (HP part number 1250-1462) as a

connector saver for R CHANNEL IN.

Agilent Technologies 8753D — Chapter 3 Making Mixer Measurements diagram (page 154)
Fig. 82 · Chapter 3 Making Mixer Measurements

Figure 3-20. Connections for the First P ortion of Conversion Compression Measurement

5. T o view the absolute input power to the analyzer's R-channel, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNN

MEAS

INPUT PORTS

R

6. T o store a trace of the receiver power versus the source power into memory and view

data/memory , press:

DISPLA Y

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DATA ! MEMORY

NNNNNNNNNNNNNNNNNNNNNNNNNN

DATA/MEM

This removes the loss between the output of the mixer and the input to the receiver , and

provides a linear power sweep for use in subsequent measurements .

7. Make the connections as shown in Figure 3-21.

Making Mixer Measurements 3-29

Show 4 more labels / short text on this page
· · ·
Page 155 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNN
Page 155 · click to enlarge
Page 155

T o prevent connector damage , use an adapter (HP part number 1250-1462) as a

connector saver for R CHANNEL IN.

Agilent Technologies 8753D — Chapter 3 Making Mixer Measurements diagram (page 155)
Fig. 83 · Chapter 3 Making Mixer Measurements

Figure 3-21.

Connections for the Second P ortion of Conversion Compression Measurement

8. T o set the frequency o set mode LO frequency , press:

SYSTEM

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INSTRUMENT MODE

FREQ OFFS MENU

NNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LO MENU

FREQUENCY:CW

M/

9. T o select the converter type , press:

NNNNNNNNNNNNNNNNNNNN

RETURN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

UP CONVERTER

10. T o select a low-side LO measurement con guration, press:

NNNNNNNNNNNNNNNNN

RF>LO

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FREQ OFFS ON

In this low-side LO , up converter measurement, the analyzer source frequency is o set lower

than the receiver frequency . The analyzer source frequency can be determined from the

following equation:

receiver frequency (800 MHz) LO frequency (600 MHz) = 200 MHz

3-30 Making Mixer Measurements

Show 4 more labels / short text on this page
· · 600 ·
Page 156 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNN
Page 156 · click to enlarge
Page 156

The measurements setup diagram is shown in Figure 3-22.

Agilent Technologies 8753D — Chapter 3 Making Mixer Measurements diagram (page 156)
Fig. 84 · Chapter 3 Making Mixer Measurements

Figure 3-22. Measurement Setup Diagram Shown on Analyzer Display

11. T o view the mixer's output power as a function of its input power , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

VIEW MEASURE

12. T o set up an active marker to search for the 1 dB compression point of the mixer , press:

SCALE REF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AUTO SCALE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER F CTN

MKR SEARCH ON

SEARCH:MAX

13. Press:

NNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER

MKR ZERO

MARKER F CTN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MKR SEARCH ON

NNNNNNNNNNNNNNNNNNNN

TARGET

The measurement results show the mixer's 1 dB compression point. By changing the target

value , you can easily locate other compression points (for example , 0.5 dB , 3 dB). See

Figure 3-22.

14. Read the compressed power on by turning marker o .

NNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER

MODE

MODE OFF .

Making Mixer Measurements 3-31

Show 13 more labels / short text on this page
· · · · · · · · 1 · · · ·
Page 157 of the Agilent Technologies 8753D user manual — Chapter 3 Making Mixer Measurements
Page 157 · click to enlarge
Page 157
Agilent Technologies 8753D — Chapter 3 Making Mixer Measurements diagram (page 157)
Fig. 85 · Chapter 3 Making Mixer Measurements

Figure 3-23.

Example Swept P ower Conversion Compression Measurement

3-32 Making Mixer Measurements

j

Page 158 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 158 · click to enlarge
Page 158

Isolation is the measure of signal leakage in a mixer . F eedthrough is speci cally the forward

signal leakage to the IF port. High isolation means that the amount of leakage or feedthrough

between the mixer's ports is very small. Isolation measurements do not use the frequency

o set mode . Figure 3-24 illustrates the signal ow in a mixer .

Agilent Technologies 8753D — Chapter 3 Making Mixer Measurements diagram (page 158)
Fig. 86 · Chapter 3 Making Mixer Measurements

Figure 3-24. Signal Flow in a Mixer Example

The RF and LO feedthrough signals may appear at the mixer IF output, together with the

desired IF signal.

The LO to RF isolation and the LO feedthrough are typically measured with the third port

terminated in 50 ohms . Measurement of the RF feedthrough is made as the LO signal is being

applied to the mixer .

LO to RF Isolation

1. Initialize the analyzer by pressing

PRESET

.

2. T o select the analyzer frequency range and source power , press:

ST ART

M/

STOP

M/

N

POWER

NNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PWR RANGE MAN

MENU

This source stimulates the mixer's LO port.

3. T o select a ratio B/R measurement, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNN

MEAS

INPUT PORTS

B/R

4. Make the connections as shown in Figure 3-25.

Making Mixer Measurements 3-33

Show 15 more labels / short text on this page
Isolation Example Measurements · · · 10 · · · · 3000 · · · · 0 · · ·
Page 159 of the Agilent Technologies 8753D user manual — SCALE REF
Page 159 · click to enlarge
Page 159
Agilent Technologies 8753D — Chapter 3 Making Mixer Measurements diagram (page 159)
Fig. 87 · Chapter 3 Making Mixer Measurements

Figure 3-25. Connections for a Response Calibration

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNN

P erform a response calibration by pressing

CAL

CALIBRATE MENU

RESPONSE

THRU .

A full 2 port calibration will increase the accuracy of isolation measurements .

Refer to Chapter 5, \Optimizing Measurement Results ."

6. Make the connections as shown in Figure 3-26.
Agilent Technologies 8753D — CAL diagram (page 159)
Fig. 88

Figure 3-26. Connections for a Mixer Isolation Measurement

7. T o adjust the display scale , press:

SCALE REF

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AUTO SCALE

The measurement results show the mixer's LO to RF isolation.

3-34 Making Mixer Measurements

· ·

Page 160 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 160 · click to enlarge
Page 160
Agilent Technologies 8753D — Chapter 3 Making Mixer Measurements diagram (page 160)
Fig. 89 · Chapter 3 Making Mixer Measurements

Figure 3-27. Example Mixer LO to RF Isolation Measurement

The procedure and equipment con guration necessary for this measurement are very similar

to those above , with the addition of an external source to drive the mixer's LO port as we

measure the mixer's RF feedthrough. RF feedthrough measurements do not use the frequency

o set mode .

1. Select the CW LO frequency and source power from the front panel of the external source .

CW frequency = 300 MHz

P ower = 10 dBm

2. Initialize the analyzer by pressing

PRESET

.

3. T o select the analyzer's frequency range and source power , press:

ST ART

M/

STOP

M/

NNNNNNNNNNNNNNNNN

POWER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PWR RANGE MAN

MENU

This signal stimulates the mixer's RF port.

4. T o select a ratio measurement, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNN

MEAS

INPUT PORTS

B/R

Isolation is dependent on LO power level and frequency . T o ensure good

test results , you should choose these parameters as close to actual operating

conditions as possible .

5. Make the connections as shown in Figure 3-28.

Making Mixer Measurements 3-35

Show 15 more labels / short text on this page
RF F eedthrough · · · 10 · · · · 3000 · · · · 0 · · ·
Page 161 of the Agilent Technologies 8753D user manual — CAL
Page 161 · click to enlarge
Page 161

Figure 3-28. Connections for a Response Calibration

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNN

P erform a response calibration by pressing

CAL

CALIBRATE MENU

RESPONSE

THRU

7. Make the connections as shown in Figure 3-29.
Agilent Technologies 8753D — CAL diagram (page 161)
Fig. 90

Figure 3-29. Connections for a Mixer RF F eedthrough Measurement

8. Connect the external LO source to the mixer's LO port.
9. The measurement results show the mixer's RF feedthrough.

Y ou may see spurious responses on the analyzer trace due to interference

caused by LO to IF leakage in the mixer . This can be reduced with averaging or

by reducing the IF bandwidth.

3-36 Making Mixer Measurements

Page 162 of the Agilent Technologies 8753D user manual — Chapter 3 Making Mixer Measurements
Page 162 · click to enlarge
Page 162
Agilent Technologies 8753D — Chapter 3 Making Mixer Measurements diagram (page 162)
Fig. 91 · Chapter 3 Making Mixer Measurements

Figure 3-30. Example Mixer RF F eedthrough Measurement

Y ou can measure the IF to RF isolation in a similar manner , but with the following

modi cations:

use the analyzer source as the IF signal drive

view the leakage signal at the RF port

Making Mixer Measurements 3-37

Page 163 of the Agilent Technologies 8753D user manual — Chapter 3 Making Mixer Measurements
Page 163 · click to enlarge
Page 163
Page 163 of the Agilent Technologies 8753D user manual — Chapter 3 Making Mixer Measurements
Page 164 of the Agilent Technologies 8753D user manual — Chapter 4 Printing, Plotting, and Saving Measurement Results
Page 164 · click to enlarge
Page 164

Printing, Plotting, and Saving Measurement

This chapter contains instructions for the following tasks:

Printing or plotting your measurement results

Con guring a print function

De ning a print function

Printing one measurement per page

Printing multiple measurements per page

Con guring a plot function

De ning a plot function

Plotting one measurement per page using a pen plotter

Plotting multiple measurements per page using a pen plotter

Plotting a measurement to disk

Outputting plot les from a PC to a plotter

Outputting plot les from a PC to an HPGL compatible printer

Outputting single page plots using a printer

Outputting multiple plots to a single page using a printer

Plotting Multiple Measurements per page from disk

Titling the displayed measurement

Con guring the analyzer to produce a time stamp

Aborting a print or plot process

Printing or plotting the list values or operating parameters

Solving problems with printing or plotting

Saving and recalling instrument states

Saving an instrument state

Re-saving an instrument state

F ormatting a disk

Solving problems with saving or recalling les

Printing, Plotting, and Saving Measurement Results 4-1

Show 7 more labels / short text on this page
Results · Printing time · Plotting time · Saving measurement results · Deleting a le · Renaming a le · Recalling a le
Page 165 of the Agilent Technologies 8753D user manual — Chapter 4 Printing, Plotting, and Saving Measurement Results
Page 165 · click to enlarge
Page 165

Where to Look for More Information

A dditional information about many of the topics discussed in this chapter is located in the

following areas:

Chapter 2, \Making Measurements ," contains step-by-step procedures for making

measurements or using particular functions .

Chapter 8, \Menu Maps ," shows softkey menu relationships .

Chapter 9, \Key De nitions ," describes all the front panel keys , softkeys , and their

corresponding HP-IB commands .

Chapter 11, \Compatible P eripherals ," lists measurement and system accessories , and other

applicable equipment compatible with the analyzers . An HP-IB programming overview is also

included.

4-2 Printing, Plotting, and Saving Measurement Results

Page 166 of the Agilent Technologies 8753D user manual — HP-IB HP 10833A/B/D
Page 166 · click to enlarge
Page 166

Printing or Plotting Y our Measurement Results

Y ou can print your measurement results to the following peripherals:

printers with HP-IB interfaces

printers with parallel interfaces

printers with serial interfaces

Y ou can plot your measurement results to the following peripherals:

HPGL compatible printers with HP-IB interfaces

HPGL compatible printers with parallel interfaces

plotters with HP-IB interfaces

plotters with parallel interfaces

plotters with serial interfaces

Refer to the \Compatible P eripherals" chapter for a list of recommended peripherals .

Con guring a Print Function

All copy con guration settings are stored in non-volatile memory . Therefore , they are not

a ected if you press

PRESET

or switch o the analyzer power .

1. Connect the printer to the interface port.

Printer Interface Recommended Cables

P arallel HP 92284A

HP-IB HP 10833A/B/D

Agilent Technologies 8753D — HP-IB HP 10833A/B/D diagram (page 166)
Fig. 92

Figure 4-1. Printer Connections to the Analyzer

Printing, Plotting, and Saving Measurement Results 4-3

Serial HP 24542G

Page 167 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 167 · click to enlarge
Page 167

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SET ADDRESSES

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRINTER PORT

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRNTR TYPE until the correct printer choice

LOCAL

appears:

NNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNN

DeskJet (except for HP DeskJet 540 and DeskJet 850C)

NNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNN

Epson-P2 (printers that conform to the ESC/P2 printer control language)

NNNNNNNNNNNNNNNNNNNN

DJ 540 (for use with the HP DeskJet 540 and DeskJet 850C)

NNNNNNNNNNNNNNNNNNNN

DJ 540 converts 100 dpi raster information to 300 dpi raster format.

If your DeskJet printer does not support the 100 dpi raster format and your

NNNNNNNNNNNNNNNNNNNN

printing results seem to be less than normal size , select

DJ 540 .

3. Select one of the following printer interfaces:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRNTR PORT HPIB if your printer has an HP-IB interface , and then con gure the

print function as follows:

a. Enter the HP-IB address of the printer , followed by

.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

b.

LOCAL

SYSTEM CONTROLLER if there is no external controller connected to

the HP-IB bus .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

c. Press

LOCAL

USE PASS CONTROL if there is an external controller connected to

the HP-IB bus .

NNNNNNNNNNNNNNNNNNNNNNNNNN

PARALLEL if your printer has a parallel (centronics) interface , and then con gure

the print function as follows:

and then select the parallel port interface function by pressing

NNNNNNNNNNNNNNNNNNNNNNNNNN

LOCAL

PARALLEL

until the correct function appears .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PARALLEL [COPY] , the parallel port is dedicated for normal copy

device use (printers or plotters).

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PARALLEL [GPIO] , the parallel port is dedicated for general purpose

I/O , and cannot be used for printing or plotting.

4-4 Printing, Plotting, and Saving Measurement Results

Show 14 more labels / short text on this page
Press · · ThinkJet (QuietJet) · LaserJet · PaintJet · Selecting · Choose · Press · and · and · Choose · Press · If you choose · If you choose
Page 168 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 168 · click to enlarge
Page 168

NNNNNNNNNNNNNNNNNNNN

SERIAL if your printer has a serial (RS-232) interface , and then con gure the print

function as follows:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRINTER BAUD RATE and enter the printer's baud rate , followed by

.

a. Press

b. T o select the transmission control method that is compatible with your printer , press

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

XMIT CNTRL (transmit control - handshaking protocol) until the correct method

appears .

NNNNNNNNNNNNNNNNNNNNNNNNNN

Xon-Xoff , the handshake method allows the printer to control the

data exchange by transmitting control characters to the network analyzer .

NNNNNNNNNNNNNNNNNNNNNNN

DTR-DSR , the handshake method allows the printer to control the data

exchange by setting the electrical voltage on one line of the RS-232 serial cable .

NNNNNNNNNNNNNNNNNNNNNNN

DTR-DSR handshake takes place in the hardware rather than the

rmware or software , it is the fastest transmission control method.

De ning a Print Function

The print de nition is set to default values whenever the power is cycled.

However , you can save the print de nition by saving the instrument state .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

1. Press

COPY

DEFINE PRINT .

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRINTMONOCHROME or

PRINT: COLOR .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRINT: MONOCHROME if you are using a black and white printer , or you want just

black and white from a color printer .

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRINT: COLOR if you are using a color printer .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AUTO-FEED until the correct choice (ON or OFF) is high-lighted.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AUTO-FEED ON if you want to print one measurement per page .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AUTO-FEED OFF if you want to print multiple measurements per page .

Laser printers and some DeskJet printers do not begin to print until a full page ,

or a partial page and a form feed, have been received.

Printing, Plotting, and Saving Measurement Results 4-5

Show 11 more labels / short text on this page
Choose · If you choose · If you choose · Because the · · Press · Choose · Choose · Press · Choose · Choose
Page 169 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 169 · click to enlarge
Page 169

If Y ou are Using a Color Printer

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRINT COLORS .

2. If you want to modify the print colors , select the print element and then choose an available

color .

Y ou can set all the print elements to black to create a hardcopy in black and

white .

Since the media color is white or clear , you could set a print element to white if

you do not want that element to appear on your hardcopy .

T o Reset the Printing P arameters to Default V alues

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DEFINE PRINT

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DEFAULT PRNT SETUP .

COPY

T able 4-1. Default V alues for Printing P arameters

Printing P arameter Default

A uto F eed ON

Channel 1 Data Magenta

Channel 1 Memory Green

Channel 2 Data Blue

Channel 2 Memory Red

Printing One Measurement P er P age

1. Con gure and de ne the print function, as explained in \Con guring a Print Function" and

\De ning a Print Function" located earlier in this chapter .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

COPY

PRINT MONOCHROME .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

If you de ned the

AUTO-FEED OFF , press

PRINTER FORM FEED after the message COPY

OUTPUT COMPLETED appears .

4-6 Printing, Plotting, and Saving Measurement Results

Show 10 more labels / short text on this page
Press · Press · · Printer Mode Monochrome · Printer Colors · Graticule Cyan · W arning Black · T ext Black · Press ·
Page 170 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 170 · click to enlarge
Page 170

Printing Multiple Measurements P er P age

1. Con gure and de ne the print function, as explained in \Con guring a Print Function" and

\De ning a Print Function" located earlier in this chapter .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

COPY

DEFINE PRINT and then press

AUTO-FEED until the softkey label appears as

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AUTO-FEED OFF .

NNNNNNNNNNNNNNNNNNNN

RETURN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRINT MONOCHROME to print a measurement on the rst half page .

4. Make the next measurement that you want to see on your hardcopy . Figure 4-2 shows an

example of a hardcopy where two measurements appear .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

COPY

PRINT MONOCHROME to print a measurement on the second half page .

This feature will not work for all printers due to di erences in printer

resolution.

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 170)
Fig. 93

Figure 4-2. Printing Two Measurements

Printing, Plotting, and Saving Measurement Results 4-7

Show 5 more labels / short text on this page
Press · · Press · Press ·
Page 171 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNN
Page 171 · click to enlarge
Page 171

Con guring a Plot Function

All copy con guration settings are stored in non-volatile memory . Therefore , they are not

a ected if you press

or switch o the analyzer power .

PRESET

1. Connect the peripheral to the interface port.

P eripheral Interface Recommended Cables

P arallel HP 92284A

HP-IB HP 10833A/33B/33D

Figure 4-3. P eripheral Connections to the Analyzer

If Y ou are Plotting to an HPGL/2 Compatible Printer

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOCAL

SET ADDRESSES

PRINTER PORT and then press

PRNTR TYPE until the correct

printer choice appears:

N

NNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNN

DeskJet (only DeskJet 1200C and DeskJet 1600C)

NNNNNNNNNNNNNNNNNNNNNNNNNN

LaserJet (only LaserJet III and IV)

NNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNN

Epson-P2 (printers that conform to the ESC/P2 printer control language)

4-8 Printing, Plotting, and Saving Measurement Results

Show 5 more labels / short text on this page
Serial HP 24542G · Press · · ThinkJet (QuietJet) · PaintJet
Page 172 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 172 · click to enlarge
Page 172
3. Con gure the analyzer for one of the following printer interfaces:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRNTR PORT HPIB if your printer has an HP-IB interface , and then con gure the

print function as follows:

a. Enter the HP-IB address of the printer (default is 01), followed by

.

b.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOCAL

SYSTEM CONTROLLER if there is no external controller connected to

the HP-IB bus .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

c. Press

LOCAL

USE PASS CONTROL if there is an external controller connected to

the HP-IB bus .

NNNNNNNNNNNNNNNNNNNNNNNNNN

PARALLEL if your printer has a parallel (centronics) interface , and then con gure

the print function as follows:

LOCAL

and then select the parallel port interface function by pressing

NNNNNNNNNNNNNNNNNNNNNNNNNN

PARALLEL

until the correct function appears .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PARALLEL [COPY] , the parallel port is dedicated for normal copy

device use (printers or plotters).

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PARALLEL [GPIO] , the parallel port is dedicated for general purpose

I/O , and cannot be used for printing or plotting.

NNNNNNNNNNNNNNNNNNNN

SERIAL if your printer has a serial (RS-232) interface , and then con gure the print

function as follows:

a. Press

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRINTER BAUD RATE and enter the printer's baud rate , followed by

.

b. T o select the transmission control method that is compatible with your printer , press

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

XMIT CNTRL (transmit control - handshaking protocol) until the correct method

appears .

NNNNNNNNNNNNNNNNNNNNNNNNNN

Xon-Xoff , the handshake method allows the printer to control the

data exchange by transmitting control characters to the network analyzer .

NNNNNNNNNNNNNNNNNNNNNNN

DTR-DSR , the handshake method allows the printer to control the data

exchange by setting the electrical voltage on one line of the RS-232 serial cable .

NNNNNNNNNNNNNNNNNNNNNNN

DTR-DSR handshake takes place in the hardware rather than the

rmware or software , it is the fastest transmission control method.

4. Press

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOCAL

SET ADDRESSES

PLOTTER PORT and then

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLTR TYPE [HPGL PRT] appears .

Printing, Plotting, and Saving Measurement Results 4-9

Show 14 more labels / short text on this page
Choose · Press · and · and · Choose · Press · If you choose · If you choose · Choose · If you choose · If you choose · Because the · · PLTR TYPE until
Page 173 of the Agilent Technologies 8753D user manual — N
Page 173 · click to enlarge
Page 173

If Y ou are Plotting to a P en Plotter

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOTTER PORT and then

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOCAL

SET ADDRESSES

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLTR TYPE [PLOTTER] appears .

2. Con gure the analyzer for one of the following plotter interfaces:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLTR PORT HPIB if your plotter has an HP-IB interface , and then con gure the

plot function as follows:

a. Enter the HP-IB address of the printer (default is 05), followed by

.

b.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SYSTEM CONTROLLER if there is no external controller connected to

LOCAL

the HP-IB bus .

c. Press

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

USE PASS CONTROL if there is an external controller connected to

LOCAL

the HP-IB bus .

NNNNNNNNNNNNNNNNNNNNNNNNNN

PARALLEL if your printer has a parallel (centronics) interface , and then con gure

the print function as follows:

NNNNNNNNNNNNNNNNNNNNNNNNNN

LOCAL

and then select the parallel port interface function by pressing

PARALLEL

until the correct function appears .

N

PARALLEL [COPY] , the parallel port is dedicated for normal copy

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

device use (printers or plotters).

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PARALLEL [GPIO] , the parallel port is dedicated for general purpose

I/O , and cannot be used for printing or plotting.

N

NNNNNNNNNNNNNNNNNNN

SERIAL if your printer has a serial (RS-232) interface , and then con gure the print

function as follows:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

a. Press

PRINTER BAUD RATE and enter the printer's baud rate , followed by

.

b. T o select the transmission control method that is compatible with your printer , press

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

XMIT CNTRL (transmit control - handshaking protocol) until the correct method

appears .

N

NNNNNNNNNNNNNNNNNNNNNNNNN

Xon-Xoff , the handshake method allows the printer to control the

data exchange by transmitting control characters to the network analyzer .

NNNNNNNNNNNNNNNNNNNNNNN

DTR-DSR , the handshake method allows the printer to control the data

exchange by setting the electrical voltage on one line of the RS-232 serial cable .

N

NNNNNNNNNNNNNNNNNNNNNN

DTR-DSR handshake takes place in the hardware rather than the

rmware or software , it is the fastest transmission control method.

4-10 Printing, Plotting, and Saving Measurement Results

Show 15 more labels / short text on this page
Press · · PLTR TYPE until · Choose · Press · and · and · Choose · Press · If you choose · If you choose · Choose · If you choose · If you choose · Because the
Page 174 of the Agilent Technologies 8753D user manual — N
Page 174 · click to enlarge
Page 174

If Y ou are Plotting to a Disk Drive

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNN

DISK .

LOCAL

SET ADDRESSES

PLOTTER PORT

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SELECT DISK and select the disk drive that you will plot to .

SA VE/RECALL

N

INTERNAL DISK if you will plot to the analyzer internal disk drive .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EXTERNAL DISK if you will plot to a disk drive that is external to the analyzer .

Then con gure the disk drive as follows:

a.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CONFIGURE EXT DISK

ADDRESSDISK and enter the HP-IB address to the disk

drive (default is 00) followed by

.

b.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DISK UNIT NUMBER and enter the drive where your disk is located,

LOCAL

.

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

If your storage disk is partitioned, press

VOLUME NUMBER and enter the volume number

where you want to store the instrument state le .

Printing, Plotting, and Saving Measurement Results 4-11

Show 10 more labels / short text on this page
Press · · Press · · Choose · Choose · Press · Press · · followed by
Page 175 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 175 · click to enlarge
Page 175

De ning a Plot Function

The plot de nition is set to default values whenever the power is cycled.

However , you can save the plot de nition by saving the instrument state .

1. Press

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DEFINE PLOT .

COPY

2. Choose which of the following measurement display elements that you want to appear on

your plot:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOT DATA ON if you want the measurement data trace to appear on your plot.

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOT MEM ON if you want the displayed memory trace to appear on your plot.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOT GRAT ON if you want the graticule and the reference line to appear on your

plot.

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOT TEXT ON if you want all of the displayed text to appear on your plot. (This

does not include the marker values or softkey labels .)

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOT MKR ON if you want the displayed markers , and marker values , to appear on

your plot.

Agilent Technologies 8753D — N diagram (page 175)
Fig. 94

Figure 4-4. Plot Components A vailable through De nition

Selecting A uto-F eed

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AUTO-FEED until the correct choice is high-lighted.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AUTO-FEED ON if you want a \page eject" sent to the plotter or HPGL compatible

NNNNNNNNNNNNNN

printer after each time you press

PLOT .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AUTO-FEED OFF if you want multiple plots on the same sheet of paper .

4-12 Printing, Plotting, and Saving Measurement Results

Show 10 more labels / short text on this page
· Choosing Display Elements · Choose · Choose · Choose · Choose · Choose · Press · Choose · Choose
Page 176 of the Agilent Technologies 8753D user manual — FFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFF
Page 176 · click to enlarge
Page 176

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AUTO-FEED ON when you are plotting to a quadrant.

Selecting P en Numbers and Colors

NNNNNNNNNNNNNN

MORE and select the plot element where you want to change the pen number . F or

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PEN NUM DATA and then modify the pen number . The pen number selects

the color if you are plotting to an HPGL/2 compatible color printer .

after each modi cation.

The following color assignments are valid for HPGL/2 compatible color printers

only . When using word processor or graphics presentation programs , di erent

colors may be assigned to the pen numbers .

T able 4-2. Default P en Numbers and Corresponding Colors

T able 4-3. Default P en Numbers for Plot Elements

Corresponding K ey Plot Element Channel 1

F

FFFFFFFFFFFFFFFFFFFFFFFFFFFFFF

PEN NUM DATA Measurement Data Trace 2 3

FFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFF

PEN NUM MEMORY Displayed Memory Trace 5 6

FFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFF

PEN NUM GRATICULE Graticule and Reference Line 1 1

F

FFFFFFFFFFFFFFFFFFFFFFFFFFFFFF

PEN NUM TEXT Displayed T ext 7 7

FFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFF

PEN NUM MARKER Displayed Markers and V alues 7 7

Y ou can set all the pen numbers to black for a plot in black and white .

Y ou must de ne the pen numbers for each measurement channel (channel 1

and channel 2).

Printing, Plotting, and Saving Measurement Results 4-13

Show 18 more labels / short text on this page
The peripheral ignores · Press · example , press · Press · P en · Color · Number · 0 white · 1 cyan · 2 magenta · 3 blue · 4 yellow · 5 green · 6 red · 7 black · Channel 2 · P en Numbers · P en Numbers
Page 177 of the Agilent Technologies 8753D user manual — N
Page 177 · click to enlarge
Page 177

NNNNNNNNNNNNNN

MORE and select each plot element line type that you want to modify .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LINE TYPE DATA to modify the line type for the data trace . Then enter the new

line type (see Figure 4-5), followed by

.

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LINE TYPE MEMORY to modify the line type for the memory trace . Then enter the

new line type (see Figure 4-5), followed by

.

T able 4-4. Default Line Types for Plot Elements

Plot Elements Channel 1

Data Trace 7 7

Memory Trace 7 7

Agilent Technologies 8753D — N diagram (page 177)
Fig. 95

Figure 4-5. Line Types A vailable

Y ou must de ne the line types for each measurement channel (channel 1 and

channel 2).

4-14 Printing, Plotting, and Saving Measurement Results

Show 7 more labels / short text on this page
Selecting Line Types · Press · Select · Select · Channel 2 · Line Type Numbers · Line Type Numbers
Page 178 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 178 · click to enlarge
Page 178

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SCALE PLOT until the selection appears that you want.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SCALE PLOT [FULL] if you want the normal scale selection for plotting. This

includes space for all display annotations such as marker values and stimulus values . The

entire analyzer display ts within the de ned boundaries of P1 and P2 on the plotter ,

while maintaining the exact same aspect ratio as the display .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SCALE PLOT [GRAT] if you want the outer limits of the graticule to correspond to

the de ned P1 and P2 scaling point on the plotter . (Intended for plotting on preprinted

rectangular or polar) forms .

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 178)
Fig. 96

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Figure 4-6. Locations of P1 and P2 in

SCALE PLOT [GRAT] Mode

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOT SPEED until the plot speed appears that you want.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOT SPEED [FAST] for normal plotting.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOT SPEED [SLOW] for plotting directly on transparencies .(The slower speed

provides a more consistent line width.)

Printing, Plotting, and Saving Measurement Results 4-15

Show 8 more labels / short text on this page
Choosing Scale · Press · Choose · Choose · Choosing Plot Speed · Press · Choose · Choose
Page 179 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 179 · click to enlarge
Page 179

T o Reset the Plotting P arameters to Default V alues

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DEFINE PLOT

NNNNNNNNNNNNNN

MORE

NNNNNNNNNNNNNN

MORE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DEFAULT PLOT SETUP .

COPY

T able 4-5. Plotting P arameter Default V alues

Plotting P arameter Default

Select Quadrant Full page

A uto F eed ON

De ne Plot All plot elements on

Plot Speed F ast

Line Type 7 (solid line)

P en NumbersChannel 1
P en NumbersChannel 2

Plotting One Measurement P er P age Using a P en Plotter

1. Con gure and de ne the plot, as explained in \Con guring a Plot Function" and \De ning a

Plot Function" located earlier in this chapter .

NNNNNNNNNNNNNN

COPY

PLOT .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

If you de ned the

AUTO-FEED OFF , press

PLOTTER FORM FEED after the message COPY

OUTPUT COMPLETED appears .

4-16 Printing, Plotting, and Saving Measurement Results

Show 15 more labels / short text on this page
Press · · Plot Scale Full · Data 2 · Memory 5 · Graticule 1 · T ext 7 · Marker 7 · Data 3 · Memory 6 · Graticule 1 · T ext 7 · Marker 7 · Press ·
Page 180 of the Agilent Technologies 8753D user manual — COPY
Page 180 · click to enlarge
Page 180

Plotting Multiple Measurements P er P age Using a P en Plotter

1. Con gure and de ne the plot, as explained in \Con guring a Plot Function" and \De ning a

Plot Function" located earlier in this chapter .

NNNNNNNNNNNNNNNNNNNNNNNNNN

COPY

SEL QUAD .

3. Choose the quadrant where you want your displayed measurement to appear on the

hardcopy . The following quadrants are available:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LEFT UPPER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LEFT LOWER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RIGHT UPPER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RIGHT LOWER

NNNNNNNNNNNNNNNNNNNNNNNNNN

The selected quadrant will appear in the brackets under

SEL QUAD .

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 180)
Fig. 97

Figure 4-7. Plot Quadrants

NNNNNNNNNNNNNN

PLOT .

5. Make the next measurement that you want to see on your hardcopy .

NNNNNNNNNNNNNNNNNNNNNNNNNN

SEL QUAD and choose another quadrant where you want to place the displayed

COPY

measurement.

7. Repeat the previous three steps until you have captured the results of up to four

measurements .

Printing, Plotting, and Saving Measurement Results 4-17

Show 5 more labels / short text on this page
Press · · Press · Press ·
Page 181 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 181 · click to enlarge
Page 181

If Y ou are Plotting to an HPGL Compatible Printer

1. Con gure and de ne the plot, as explained in \Con guring a Plot Function" and \De ning a

Plot Function" located earlier in this chapter .

N

PLOT

NNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOTTER FORM FEED to print the data the printer has received.

COPY

Use test sequencing to automatically plot all four S-parameters .

1. Set all measurement parameters .
2. P erform a full 2-port calibration.
3. Enter the test sequence:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

NEW SEQ/MODIFY SEQ

SEQUENCE 1 SEQ1

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MEAS

ReflFWD S11 (A/R)

NNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNN

COPY

SEL QUAD

LEFT UPPER

PLOT

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MEAS

TransFWD S21 (B/R)

NNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNN

COPY

SEL QUAD

LEFT LOWER

PLOT

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ReflREV S22 (B/R)

MEAS

NNNNNNNNNNNNNNNNNNNNNNNNNN

SEL QUAD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RIGHT UPPER

NNNNNNNNNNNNNN

PLOT

COPY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TransREV S12 (B/R)

MEAS

COPY

NNNNNNNNNNNNNNNNNNNNNNNNNN

SEL QUAD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RIGHT LOWER

NNNNNNNNNNNNNN

PLOT

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

DONE SEQ MODIFY

4. Run the test sequence by pressing:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

DO SEQUENCE

SEQUENCE 1 SEQ1

4-18 Printing, Plotting, and Saving Measurement Results

Show 24 more labels / short text on this page
Press · · · · · · · · · · · · · · · · · · · · · · ·
Page 182 of the Agilent Technologies 8753D user manual — COPY
Page 182 · click to enlarge
Page 182

Plotting a Measurement to Disk

The plot les that you generate from the analyzer , contain the HPGL representation of the

measurement display . The les will not contain any setup or formfeed commands .

1. Con gure the analyzer to plot to disk.

a.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNN

DISK .

LOCAL

SET ADDRESSES

PLOTTER PORT

b.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SELECT DISK and select the disk drive that you will plot to .

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INTERNAL DISK if you will plot to the analyzer internal disk drive .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EXTERNAL DISK if you will plot to a disk drive that is external to the analyzer .

Then con gure the disk drive as follows:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CONFIGURE EXT DISK

ADDRESSDISK and enter the HP-IB address to the

disk drive (default is 00) followed by

.

N

DISK UNIT NUMBER and enter the drive where your disk is located,

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOCAL

.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

If your storage disk is partitioned, press

VOLUME NUMBER and enter the volume

number where you want to store the instrument state le .

NNNNNNNNNNNNNN

PLOT .

COPY

The analyzer assigns the rst available default lename for the displayed directory . F or

example , the analyzer would assign PLOT00FP for a LIF format (PLOT00.FP for a DOS

format) if there were no previous plot les saved to the disk.

The gure below shows the three parts of the le name that is generated automatically by

the analyzer whenever a plot is requested. The two digit sequence number is incremented

by one each time a le with a default name is added to the directory .

Agilent Technologies 8753D — COPY diagram (page 182)
Fig. 98

Figure 4-8. A utomatic File Naming Convention for LIF F ormat

Printing, Plotting, and Saving Measurement Results 4-19

Show 12 more labels / short text on this page
Press · · Press · · Choose · Choose · Press · Press · · followed by · Press ·
Page 183 of the Agilent Technologies 8753D user manual — R
Page 183 · click to enlarge
Page 183

T o Output the Plot Files

Y ou can plot the les to a plotter from a personal computer .

Y ou can output your plot les to an HPGL compatible printer , by following the sequence

in \Outputting Plot Files from a PC to an HPGL Compatible Printer" located later in this

chapter .

Y ou can run a program that plots all of the les , with the root lename of PLOT , to an

HPGL compatible printer . This program is provided on the \Example Programs Disk" that is

included in the HP 8753D Network Analyzer Programmer's Guide. However , this program is

for use with LIF formatted disks and is written in HP B ASIC.

T o View Plot Files on a PC

Plot les can be viewed and manipulated on a PC using a word processor or graphics

presentation program. Plot les contain a text stream of HPGL (Hewlett-P ackard Graphics

Language) commands . In order to import a plot le into an application, that application must

have an import lter for HPGL (often times call HGL). Two such applications from the Lotus

R

suite of products are the word processor \AmiPro" and the graphics presentation package

\Freelance Graphics ."

Lotus applications are not supported by Hewlett-P ackard. The following

procedures are provided for informational use only . Other applications or other

versions of the same application may function di erently .

When viewed in such programs , the color and font size of the plot may vary from the output

of an HPGL/2 compatible color printer . The following table shows the di erences between

the color assignments of HPGL/2 compatible printers and Lotus applications . Also refer to

\Selecting P en Numbers and Colors" located earlier in this chapter .

HPGL/2 Printer Lotus Applications

No .

No .

3 blue 3 green

4 yellow 4 yellow

5 green 5 blue

6 red 6 red-violet

7 black 7 aqua

T o modify the color or font size , consult the documentation for the particular application being

used.

4-20 Printing, Plotting, and Saving Measurement Results

Show 12 more labels / short text on this page
P en · Color P en · Color · 0 white · 1 cyan · 1 black · (aqua) · 2 magenta · 2 red · (red-violet) · (magenta) · (cyan)
Page 184 of the Agilent Technologies 8753D user manual — Chapter 4 Printing, Plotting, and Saving Measurement Results
Page 184 · click to enlarge
Page 184

T o view plot les in AmiPro , perform the following steps:

1. From the FILE pull-down menu, select IMPORT PICTURE.
2. In the dialog box, change the File Type selection to HPGL. This automatically changes the

le sux in the lename box to *.PL T .

The network analyzer does not use the sux *.PL T , so you may want to change

the lename lter to *.* or some other pattern that will allow you to locate the

les you wish to import.

3. Click OK to import the le .
4. The next dialog box allows you to select paper type , rotation (landscape or portrait), and pen

colors . Y ou will probably need to change pen colors .

The network analyzer uses pen 7 for text. The default color in Ami Pro for pen

7 is aqua, which is not very readable against the typical white background. Y ou

may want to change pen 7 to black.

5. After all selections have been made , the le is imported and rendered in a small graphics

frame which can be sized to the page by grabbing one of the nodes and stretching the box as

required.

Y ou will notice that the annotation around the display is not optimum, as the Ami Pro

lter does not accurately import the HPGL command to render text.

Printing, Plotting, and Saving Measurement Results 4-21

Using AmiPro

Page 185 of the Agilent Technologies 8753D user manual — C:> TYPE PLOT00.FP > COM1
Page 185 · click to enlarge
Page 185

T o view plot les in Freelance , perform the following steps:

1. From the FILE pull-down menu, select IMPORT .
2. Set the le type in the dialog box to HGL.

The network analyzer does not use the sux *.HGL, so you may want to

change the lename lter to *.* or some other pattern that will allow you to

locate the les you wish to import.

3. Click OK to import the le .

Y ou will notice that when the trace is displayed, the text annotation will be illegible . Y ou

can easily x this with the following steps:

a. From the TEXT pull-down menu select FONT .

b. Select the type face and size . (F ourteen point text is a good place to start.)

c. Click OK to resize the font.

If you wish to modify the color of the displayed text, perform the following steps:

a. From the ARRANGE pull-down menu select UNGROUP .

b. Highlight a piece of text.

c. From the STYLE pull-down menu select A TTRIBUTES.

d. Select the desired text color and click OK.

e. Repeat steps b through d for each piece of text.

Outputting Plot Files from a PC to a Plotter

1. Connect the plotter to an output port of the computer (for example , COM1).
2. If using the COM1 port, output the le to the plotter by using the following command:

C:> TYPE PLOT00.FP > COM1

4-22 Printing, Plotting, and Saving Measurement Results

Using Freelance

Page 186 of the Agilent Technologies 8753D user manual — Chapter 4 Printing, Plotting, and Saving Measurement Results
Page 186 · click to enlarge
Page 186

Outputting Plot Files from a PC to an HPGL Compatible Printer

T o output the plot les to an HPGL compatible printer , you can use the HPGL initialization

sequence linked in a series as follows:

1. Store the HPGL initialization sequence in a le named hpglinit .
2. Store the exit HPGL mode and form feed sequence in a le named exithpgl .
3. Send the HPGL initialization sequence to the printer .
4. Send the plot le to the printer .
5. Send the exit HPGL mode and form feed sequence to the printer .
1. Store the HPGL initialization sequence .
1. Create a test le , by typing in each character as shown in the left hand column of T able 4-6.

Do not insert spaces or linefeeds . Most editors allow the inclusion of escape sequences .

F or example , in the MS-DOS editor (DOS 5.0 or greater), press CNTRL-P (hold down the

CTRL key and press P) followed by the ESCape key to create the escape character .

2. Name the le hpglinit .

T able 4-6. HPGL Initialization Commands

E conditional page eject

&12A page size 8.5 x 11

&l1O landscape orientation

(lower case l, one , capital O)

&a0L no left margin

(a, zero , capitol L)

&a400M no right margin

(a, 4, zero , zero , capitol M)

&l0E no top margin

(lower case l, zero , capitol E)

*c7680x5650Y frame size 10.66\x 7.847"

*p50x50Y move cursor to anchor point

*c0T set picture frame anchor point

*r-3U set CMY palette

%1B enter HPGL mode; cursor at PCL

As shown in T able 4-6 , the is the symbol used for the escape character ,

decimal value 27.

Printing, Plotting, and Saving Measurement Results 4-23

Command Remark · (720 decipoints/inch)

Page 187 of the Agilent Technologies 8753D user manual — Chapter 4 Printing, Plotting, and Saving Measurement Results
Page 187 · click to enlarge
Page 187
2. Store the exit HPGL mode and form feed sequence .
1. Create a test le by typing in each character as shown in the left hand column of T able 4-7 .

Do not insert spaces or linefeeds .

2. Name the le exithpgl .

T able 4-7. HPGL T est File Commands

%0A exit HPGL mode

E form feed

3. Send the HPGL initialization sequence to the printer .
4. Send the plot le to the printer .
5. Send the exit HPGL mode and form feed sequence to the printer .

Outputting Single P age Plots Using a Printer

Y ou can output plot les to an HPGL compatible printer , using the DOS command line and the

les created in the previous steps . This example assumes that the escape sequence les and the

plot les are in the current directory and the selected printer port is PRN.

C:> type hpglinit > PRN

C:> type PLOT00.FP > PRN

C:> type exithpgl > PRN

4-24 Printing, Plotting, and Saving Measurement Results

Command Remark · Command Remark

Page 188 of the Agilent Technologies 8753D user manual — PLOT00.LL
Page 188 · click to enlarge
Page 188

Outputting Multiple Plots to a Single P age Using a Printer

Refer to the \Plotting Multiple Measurements P er P age Using a Disk Drive ," located earlier in

this chapter , for the naming conventions for plot les that you want printed on the same page .

Y ou can use the following batch le to automate the plot le printing. This batch le must be

saved as \do plot.bat."

rem Namedo plot

rem Description:

rem output HPGL initialization sequence to a le:spooler

rem append all the requested plot les to the spooler

rem append the formfeed sequence to the spooler

rem copy the le to the printer

rem (This routine uses COPY instead of PRINT because COPY

rem will not return until the action is completed. PRINT

rem will queue the le so the subsequent DEL will likely

rem generate an error . COPY avoids this .)

type hpglinit > spooler

for %%i in (%1) do type %%i >> spooler

type exithpgl >> spooler

F or example , you have the following list of les to plot:

PLOT00.LL

PLOT00.LU

PLOT00.RL

PLOT00.RU

Y ou would invoke the batch print as follows:

C:> do_plot PLOT00.*

Printing, Plotting, and Saving Measurement Results 4-25

Show 9 more labels / short text on this page
rem · rem · rem · rem · rem · echo o · copy spooler LPT1 · del spooler · echo on
Page 189 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 189 · click to enlarge
Page 189

Plotting Multiple Measurements P er P age From Disk

The following procedures show you how to store plot les on a LIF formatted disk. A naming

convention is used so you can later run an HP B ASIC program on an external controller that

will output the les to the following peripherals:

a plotter with auto-feed capability , such as the HP 7550B

an HP-GL/2 compatible printer , such as the LaserJet 4 series (monochrome) or the DeskJet

1200C or DeskJet 1600C (color)

The program is contained on the \Example Programs Disk" that is provided with the

HP 8753D Network Analyzer Programmer's Guide. The le naming convention allows the

program to initiate the following:

to initialize the printer for HP-GL/2 at the beginning of a page

to plot multiple plot les on the same page

to send a page eject (form feed) to the hardcopy device , when all plots to the same page have

The plot le name is made up of four parts , the rst three are generated automatically by the

analyzer whenever a plot is requested. The two digit sequence number is incremented by one

each time a le with a default name is added to the directory .

Agilent Technologies 8753D — Chapter 4 Printing, Plotting, and Saving Measurement Results diagram (page 189)
Fig. 99 · Chapter 4 Printing, Plotting, and Saving Measurement…

Figure 4-9. Plot Filename Convention

T o Plot Multiple Measurements on a Full P age

Y ou may want to plot various les to the same page , for example , to show measurement data

traces for di erent input settings , or parameters , on the same graticule .

1. De ne the plot, as explained in \De ning the Plot Function" located earlier in this chapter .

NNNNNNNNNNNNNN

COPY

PLOT . The analyzer assigns the rst available default lename for the

displayed directory . F or example , the analyzer would assign PLOT00FP if there were no

previous plot les on the disk.

3. Press

SA VE/RECALL

and turn the front panel knob to high-light the name of the le that

you just saved.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FILE UTILITIES

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RENAME FILE and turn the front panel knob to place the " pointer

to the A character .

4-26 Printing, Plotting, and Saving Measurement Results

Show 4 more labels / short text on this page
been completed · Press · · Press
Page 190 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 190 · click to enlarge
Page 190
5. Press

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SELECT LETTER

NNNNNNNNNNNNNN

DONE .

6. De ne the next measurement plot that you will be saving to disk.

F or example , you may want only the data trace to appear on the second plot for

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

measurement comparison. In this case , you would press

COPY

DEFINE PLOT and choose

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOT DATA ON

PLOT MEM OFF

PLOT GRAT OFF

PLOT TEXT OFF

PLOT MKR OFF .

NNNNNNNNNNNNNN

COPY

PLOT . The analyzer will assign PLOT00FP because you renamed the last le

saved.

8. Press

SA VE/RECALL

and turn the front panel knob to high-light the name of the le that

you just saved.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FILE UTILITIES

RENAME FILE and turn the front panel knob to place the " pointer

to the B character .

10. Press

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SELECT LETTER

NNNNNNNNNNNNNN

DONE .

11. Continue de ning plots and renaming the saved le until you have saved all the data

that you want to put on the same page . Renaming the les as shown below allows you to

use the provided program, that organizes and plots the les , according to the le naming

convention.

Plot File Recognized Filename

First File Saved PLOT00FP A

Second File Saved PLOT00FPB

Third File Saved PLOT00FPC

F ourth File Saved PLOT00FPD

The gure below shows plots for both the frequency and time domain responses of the same

device .

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 190)
Fig. 100

Figure 4-10. Plotting Two Files on the Same P age

Printing, Plotting, and Saving Measurement Results 4-27

Show 4 more labels / short text on this page
· Press · · Press
Page 191 of the Agilent Technologies 8753D user manual — SA VE/RECALL
Page 191 · click to enlarge
Page 191

T o Plot Measurements in P age Quadrants

1. De ne the plot, as explained in \De ning the Plot Function" located earlier in this chapter .

NNNNNNNNNNNNNNNNNNNNNNNNNN

SEL QUAD .

COPY

3. Choose the quadrant where you want your displayed measurement to appear on the

NNNNNNNNNNNNNNNNNNNNNNNNNN

hardcopy . The selected quadrant appears in the brackets under

SEL QUAD .

Figure 4-11. Plot Quadrants

NNNNNNNNNNNNNN

PLOT . The analyzer assigns the rst available default lename for the selected

quadrant. F or example , the analyzer would assign PLOT01LU if there were no other left

upper quadrant plots on the disk.

5. Make the next measurement that you want to see on your hardcopy .
6. Repeat this procedure for the remaining plot les that you want to see as quadrants on a

page . If you want to see what quadrants you have already saved, press

SA VE/RECALL

view the directory .

4-28 Printing, Plotting, and Saving Measurement Results

Show 4 more labels / short text on this page
Press · · Press · to
Page 192 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNN
Page 192 · click to enlarge
Page 192

Titling the Displayed Measurement

Y ou can create a title that is printed or plotted with your measurement result.

NNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNN

DISPLA Y

MORE

TITLE to access the title menu.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ERASE TITLE .

3. Turn the front panel knob to move the arrow pointer to the rst character of the title .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SELECT LETTER .

5. Repeat the previous two steps to enter the rest of the characters in your title . Y ou can enter

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

a title that has a maximum of 50 characters . Press

BACK SPACE if you enter an incorrect

character .

NNNNNNNNNNNNNN

6. Press

DONE to complete the title entry .

Titles may also be entered using the optional external keyboard.

NNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNN

FORMFEED keys are not intended for creating display titles .

Those keys are for creating commands to send to peripherals during a sequence

program.

Printing, Plotting, and Saving Measurement Results 4-29

Show 6 more labels / short text on this page
Press · · Press · Press · The · NEWLINE and
Page 193 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 193 · click to enlarge
Page 193

Con guring the Analyzer to Produce a Time Stamp

Y ou can set a clock, and then activate it, if you want the time and date to appear on your

hardcopies .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SYSTEM

SET CLOCK .

NNNNNNNNNNNNNNNNNNNNNNNNNN

SET YEAR and enter the current year (four digits), followed by

.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SET MONTH and enter the current month of the year , followed

.

NNNNNNNNNNNNNNNNNNNNNNN

SET DAY and enter the current day of the month, followed by

.

NNNNNNNNNNNNNNNNNNNNNNNNNN

SET HOUR and enter the current hour of the day (0-23), followed by

.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SET MINUTES and enter the next immediate minute , followed by

.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ROUND SECONDS when the current time is exactly as you have set it.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TIME STAMP ON appears on the softkey label.

Aborting a Print or Plot Process

1. Press the

key to stop all data transfer .

LOCAL

2. If your peripheral is not responding, press

again or reset the peripheral.

LOCAL

Printing or Plotting the List V alues or Operating P arameters

COPY

NNNNNNNNNNNNNN

LIST and select the information that you want to appear on your hardcopy .

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LIST VALUES if you want a tabular listing of the measured data points , and

their current values , to appear on your hardcopy . This list will also include the limit test

information, if you have the limits function activated.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

OP PARMS (MKRS etc) if you want a tabular listing of the parameters for both

measurement channels to appear on your hardcopy . The parameters include: operating

parameters , marker parameters , and system parameters that relate to the control of

peripheral devices .

If Y ou want a Single P age of V alues

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRINT MONOCHROME for a printer or

NNNNNNNNNNNNNN

PLOT for a plotter peripheral, to create a

hardcopy of the displayed page of listed values .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NEXT PAGE to display the next page of listed values . Press

PREVIOUS PAGE to display

the previous page of listed values . Or , you can press

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PREVIOUS PAGE

repeatedly to display a particular page of listed values that you want to appear on your

hardcopy . Then repeat the previous step to create the hardcopy .

4-30 Printing, Plotting, and Saving Measurement Results

Show 17 more labels / short text on this page
Press · · Press · Press · Press · Press · Press · Press · Press · TIME STAMP until · Press · · Choose · Choose · Choose · Press · NEXT PAGE or
Page 194 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNN
Page 194 · click to enlarge
Page 194
3. Repeat the previous two steps until you have created hardcopies for all the desired pages of

listed values .

If you are printing the list of measurement data points , each page contains 30 lines of data.

The number of pages is determined by the number of measurement points that you have

selected under the

MENU

key .

If Y ou W ant the Entire List of V alues

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRINT ALL to print all pages of the listed values .

If you are printing the list of operating parameters , only the rst four pages are

printed. The fth page , system parameters , is printed by displaying that page

NNNNNNNNNNNNNNNNN

PRINT .

Printing, Plotting, and Saving Measurement Results 4-31

Choose · and then pressing

Page 195 of the Agilent Technologies 8753D user manual — LOCAL
Page 195 · click to enlarge
Page 195

Solving Problems with Printing or Plotting

If you encounter a problem when you are printing or plotting, check the following list for

possible causes:

Look in the analyzer display message area. The analyzer may show a message that will

identify the problem. Refer to the \Error Messages" chapter if a message appears .

If necessary , refer to the con guration procedures in this chapter to check that you have

done the following:

connected an interface cable between the peripheral and the analyzer

connected the peripheral to ac power

switched on the power

switched the peripheral on line

selected the correct printer or plotter type

If you are using a laser printer for plotting, and the printer is outputting partial plots , the

printer may require more memory and/or the page protection activated.

Consult your printer manual for information on upgrading memory and how to

activate page protection.

Make sure that the analyzer address setting for the peripheral corresponds to the actual

HP-IB address of the peripheral. The procedure is explained earlier in this chapter .

Make sure that the analyzer is in system controller mode , by pressing

LOCAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SYSTEM CONTROLLER , if the analyzer is not connected to an external controller . Otherwise ,

the analyzer must be in the pass control mode .

Substitute the interface cable .

Substitute a di erent printer or plotter .

4-32 Printing, Plotting, and Saving Measurement Results

Page 196 of the Agilent Technologies 8753D user manual — Chapter 4 Printing, Plotting, and Saving Measurement Results
Page 196 · click to enlarge
Page 196

Saving and Recalling Instrument States

Places Where Y ou Can Save

oppy disk using the analyzer's internal disk drive

oppy disk using an external disk drive

IBM compatible personal computer using HP-IB mnemonics

What Y ou Can Save to the Analyzer's Internal Memory

The number of registers that the analyzer allows you to save depends on the size of associated

error-correction sets , and memory traces . Refer to the \Preset State and Memory Allocation"

chapter for further information.

Y ou can save instrument states in the analyzer internal memory , along with the following list of

analyzer settings . The default lenames are REG.

error-corrections on channels 1 and 2

When the ac line power is switched o , the internal non-volatile memory is

retained by a battery . The data retention time with the 3 V , 1.2 Ah battery is

as follows:

T emperature at 70

C : 250 days (0.68 year) characteristically

T emperature at 40

C : 1244 days (3.4 years) characteristically

T emperature at 25

C : 10 years characteristically

What Y ou Can Save to a Floppy Disk

Y ou can save an instrument state and/or measurement results to a disk. The default lenames

are FILEn, where n gets incremented by one each time a le with a default name is added to

the directory . The default lenames for data-only les are D A T AnDn (D A T An.Dn for DOS),

where the rst n is incremented by one each time a le with a default name is added to the

directory . The second n is the channel where the measurement was made . When you save a

le to disk, you can choose to save some or all of the following:

all settings listed above for internal memory

active error-correction for the active channel only

displayed measurement data trace

Printing, Plotting, and Saving Measurement Results 4-33

Show 16 more labels / short text on this page
analyzer internal memory · displayed memory trace · print/plot de nitions · measurement setup · frequency range · number of points · sweep time · output power · sweep type · measurement parameter · · · · displayed user graphics · data only · HPGL plots
Page 197 of the Agilent Technologies 8753D user manual — Chapter 4 Printing, Plotting, and Saving Measurement Results
Page 197 · click to enlarge
Page 197

What Y ou Can Save to a Computer

Instrument states can be saved to and recalled from an external computer (system controller)

using HP-IB mnemonics . F or more information about the speci c analyzer settings that can be

saved, refer to the output commands located in the \Command Reference" chapter of the

HP 8753D Network Analyzer Programmer's Guide. F or an example program, refer to \Saving

and Recalling Instruments States" in the \Programming Examples" chapter of the

HP 8753D Network Analyzer Programmer's Guide.

4-34 Printing, Plotting, and Saving Measurement Results

Page 198 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 198 · click to enlarge
Page 198

Saving an Instrument State

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SA VE/RECALL

SELECT DISK and select one of the storage devices:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INTERNAL MEMORY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INTERNAL DISK

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EXTERNAL DISK and then con gure as follows:

a. Connect an external disk drive to the analyzer's HP-IB connector , and con gure as

follows:

b. Press

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DISK UNIT NUMBER and enter the drive where your disk is located,

LOCAL

.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

If your storage disk is partitioned, press

VOLUME NUMBER and enter the volume number

where you want to store the instrument state le .

d.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SET ADDRESSES

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ADDRESS: DISK .

e. Enter the HP-IB address of the peripheral, if the default address is incorrect

(default = 00). F ollow the entry by pressing

.

f. Press

LOCAL

and select one of the following:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SYSTEM CONTROLLER to allow the analyzer to control peripherals directly .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TALKER/LISTENER to allow the computer controller to be involved in all

peripheral access operations .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

USE PASS CONTROL to allow yourself to control the analyzer over HP-IB and

also allows the analyzer to take or pass control.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SA VE/RECALL

SAVE STATE .

The analyzer saves the state in the next available register , if you are saving to internal

memory , or saves the state to disk. Although one le is shown to represent an instrument

state on the analyzer display , each instrument state is composed of numerous les (which

can be viewed on a PC).

If you have saved enough les that you have used all the default names

(FILE00 - FILE31 for disk les , or REG1 - REG31 for memory les), you must

do one of the following in order to save more states:

rename an existing le to make a default name available

delete an existing le/register

Printing, Plotting, and Saving Measurement Results 4-35

Show 12 more labels / short text on this page
Press · · · followed by · Press · Choose · Choose · Choose · Press · · use another disk · re-save a le/register
Page 199 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 199 · click to enlarge
Page 199

Instrument states combined with measurements results can only be saved to disk. Files that

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

contain data-only , and the various save options available under the

DEFINE DISK SAVE key ,

are also only valid for disk saves .

The analyzer stores data in arrays along the processing ow of numerical data, from IF

detection to display . These arrays are points in the ow path where data is accessible , usually

via HP-IB . Y ou can choose from three di erent arrays which vary in modi cation exibility

when they are recalled.

data (raw data with error-correction applied)

format (data processed to the display format)

If you choose to save the raw data array , you will have the most exibility in modifying the

recalled measurement (including the ability to view all four S-parameters). This is because the

raw data array has the least amount of processing associated with it. Conversely , if you choose

to save the format array , your modi cation of the recalled measurement will be limited by all

the processes that are associated with that measurement result. However , the format array is

appropriate if you want to retrieve data traces that look like the currently displayed data.

De ne Save Modi cation Flexibility

Raw Data Array Most

F ormat Array Least

Y ou can also save data-only . This is saved to disk with default lenames D A T A00D1 to

D A T A31D1, for channel 1, or D A T A00D2 to D A T A31D2, for channel 2. However , these les are

not instrument states and cannot be recalled.

4-36 Printing, Plotting, and Saving Measurement Results

Show 4 more labels / short text on this page
Saving Measurement Results · raw data · During Recall · Data Array Medium
Page 200 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 200 · click to enlarge
Page 200
Agilent Technologies 8753D — Chapter 4 Printing, Plotting, and Saving Measurement Results diagram (page 200)
Fig. 101 · Chapter 4 Printing, Plotting, and Saving Measurement…

Figure 4-12. Data Processing Flow Diagram

If the analyzer has an active two-port measurement calibration, all

four S-parameters will be saved with the measurement results . All four

S-parameters may be viewed if the raw data array has been saved.

1. If you want to title the displayed measurement, refer to \Titling the Displayed

Measurement," located earlier in this chapter .

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SA VE/RECALL

SELECT DISK .

3. Choose one of the following disk drives:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INTERNAL DISK

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EXTERNAL DISK and then con gure as follows:

a. Connect an external disk drive to the analyzer's HP-IB connector , and con gure as

follows:

b. Press

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOCAL

DISK UNIT NUMBER and enter the drive where your disk is located,

.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

If your storage disk is partitioned, press

VOLUME NUMBER and enter the volume number

where you want to store the instrument state le .

d.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SET ADDRESSES

ADDRESS: DISK .

e. Enter the HP-IB address of the peripheral, if the default address is incorrect

(default = 00). F ollow the entry by pressing

.

Printing, Plotting, and Saving Measurement Results 4-37

Show 5 more labels / short text on this page
Press · · · followed by · Press
Page 201 of the Agilent Technologies 8753D user manual — N
Page 201 · click to enlarge
Page 201

f. Press

LOCAL

and select one of the following:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SYSTEM CONTROLLER to allow the analyzer to control peripherals directly .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TALKER/LISTENER to allow the computer controller to be involved in all

peripheral access operations .

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PASS CONTROL to allow yourself to control the analyzer over HP-IB and also

allows the analyzer to take or pass control.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SA VE/RECALL

DEFINE DISK-SAVE .

5. De ne the save by selecting one of the following choices:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DATA ARRAY ON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RAW ARRAY ON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FORMAT ARRAY ON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GRAPHICS ON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DATA ONLY ON (see note below)

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DATA ARRAY ON ,

RAW ARRAY ON , or

FORMAT ARRAY ON , the data is stored

to disk in IEEE-64 bit real format (for LIF disks), and 32 bit PC format for DOS disks . This

makes the DOS data les half the size of the LIF les .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DATA ARRAY ON may store data to disk in the S2P ASCII data format.

See \ASCII Data F ormats ."

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GRAPHICS ON , the user graphics area is saved. (Refer to the

HP 8753D Network Analyzer Programmer's Guide for information on using display

graphics .) The measurement display is not saved with this selection. (Refer to the

information located earlier in this chapter for a procedure that shows you how to plot

measurement displays to disk.)

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DATA ONLY ON will override all of the other save options . Because

this type of data is only intended for computer manipulation, the le contents

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DATA ONLY ON save cannot be recalled and displayed on the analyzer .

6. Choose the type of format you want:

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SAVE USING BINARY for all applications except CITI le , S2P , or CAE applications .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SAVE USING ASCII for CITI le , S2P , and CAE applications or when you want to

import the information into a spread sheet format.

N

NNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RETURN

SAVE STATE .

4-38 Printing, Plotting, and Saving Measurement Results

Show 13 more labels / short text on this page
Choose · Choose · Choose · Press · · If you select · Selecting · If you select · Selecting · of a · Choose · Choose · Press
Page 202 of the Agilent Technologies 8753D user manual — DATA ONLY ON
Page 202 · click to enlarge
Page 202

ASCII Data F ormats

CITI le (Common Instrumentation Transfer and Interchange le) is an ASCII data format

that is useful when exchanging data between di erent computers and instruments . F or more

information on the CITI le data format as well as a list of CITI le keywords , refer to Appendix

A, \The CITI le Data F ormat and Keyword Reference ."

S2P Data F ormat

Component data les contain small signal S-parameters described by frequency dependent

linear network parameters for 2 port components . These les are assigned a lename with the

sux S1 or S2 depending on which measurement channel generated the data. These les are

output only . (They cannot be read in by the analyzer .)

An S2P le is output only when the user has either a full 2-port or TRL 2-port error correction

turned on, and has selected it under

SA VE/RECALL

:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DEFINE DISK-SAVE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DATA ARRAY ON or

DATA ONLY ON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SAVE USING ASCII

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SAVE STATE

The template for component data les is as follows:

#

! indicates that all following on this line is a comment

# indicates that entries following on this line are parameters that are being

GHz, MHz, kHz, Hz

parameter S for S-parameters

format DB for dB magnitude and angle in degrees

MA for linear magnitude and angle in degrees

RI for real and imaginary pair

Rn the reference impedance in ohms for the analyzer making the measurement

(R 50 or R 75)

The \format" above is selected by the current selection under the FORMA T menu. T o select the

DB format, the FORMA T must be LOG MA G. F or MA, the FORMA T must be LIN MA G, and all

other FORMA T selections will output RI data. The S2P data will always represent the format

array data, including e ects of electrical delay and port extensions . A CITI le will be saved at

the same time . T o be consistent with previous versions , the CITI le data saved will represent

the D A T A array (corrected data) without e ects of electrical delay or port extensions .

Printing, Plotting, and Saving Measurement Results 4-39

Show 8 more labels / short text on this page
CITI le · ! comment line · <data line> · <data line> · where · speci ed · frequency · units
Page 203 of the Agilent Technologies 8753D user manual — # HZ S DB R 50
Page 203 · click to enlarge
Page 203

Here is an S2P example le for an 21 point measurement of a 20 dB attenuator:

! Network Analyzer HP 8753D.06.11 Serial No. US31240052

!

! 23 May 1997 15:26:54

# HZ S DB R 50

50000000 56.404 145.38 .0083 .3337 .0079 .1606 58.034 5.0084

1050000000 68.761 65.356 .0142 .0137 .0042 .1043 64.085 83.573

2050000000 64.108 41.723 .0253 .0068 .0147 .1675 61.954 173.75

3050000000 60.125 119.38 .0358 0.0 .0279 .1455 60.338 56.346

4050000000 61.224 32.686 .0474 .0137 .0384 .1249 61.743 169.73

5050000000 59.429 38.486 .0596 .0494 .0448 .0700 55.876 156.44

6050000000 56.035 70.648 .0681 .0975 .0553 .0315 63.449 39.47

7050000000 54.229 88.746 .0749 .1139 .0633 .0068 55.804 30.247

8050000000 61.411 111.97 .0802 .1977 .0712 .0521 51.102 97.546

9050000000 52.49 103.21 .0828 .2952 .0764 .1249 52.406 126.36

10050000000 64.291 35.461 .0875 .3213 .0775 .2252 59.417 85.038

11050000000 52.096 46.505 .0918 .4298 .0770 .2774 48.868 68.46

12050000000 49.648 78.573 .0878 .5232 .0787 .3364 50.699 77.157

13050000000 48.431 25.793 .0805 .5616 .0751 .4229 48.461 60.445

14050000000 45.984 36.612 .0717 .6097 .0651 .4202 44.971 37.711

15050000000 52.703 9.3823 .0748 .6001 .0614 .3749 46.822 23.128

16050000000 50.548 63.182 .0863 .5685 .0849 .3364 53.049 11.283

17050000000 57.776 19.931 .0973 .5877 .0971 .4229 48.105 34.254

18050000000 56.256 98.687 .1022 .7045 .0993 .5081 54.446 67.992

19050000000 76.33 149.78 .0965 .7635 .1004 .5644 48.489 48.591

20050000000 59.269 163.78 .1050 .7951 .1078 .6083 44.865 2.8304

4-40 Printing, Plotting, and Saving Measurement Results

Page 204 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 204 · click to enlarge
Page 204

Re-Saving an Instrument State

If you re-save a le , the analyzer overwrites the existing le contents .

Y ou cannot re-save a le that contains data only . Y ou must create a new le .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

1. Press

SA VE/RECALL

SELECT DISK and select the storage device .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INTERNAL MEMORY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INTERNAL DISK

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EXTERNAL DISK

NNNNNNNNNNNNNNNNNNNN

RETURN and then use the

* keys or the front panel knob to high-light the name

of the le that you want to re-save .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNN

RE-SAVE STATE

YES .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SA VE/RECALL

SELECT DISK .

2. Choose from the following storage devices:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INTERNAL MEMORY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INTERNAL DISK

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EXTERNAL DISK

NNNNNNNNNNNNNNNNNNNN

RETURN .

T o Delete an Instrument State File

* keys or the front panel knob to high-light the name of the le that you want

to delete .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FILE UTILITIES

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DELETE FILE

NNNNNNNNNNN

YES to delete all of the les that make up the

selected instrument state .

T o Delete all Files

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNN

FILE UTILITIES

DELETE ALL FILES

YES to delete all of the les that are on the

selected storage device .

Printing, Plotting, and Saving Measurement Results 4-41

Show 14 more labels / short text on this page
· Press · + · · Press · Deleting a File · Press · · Press · Press the · + · · Press · Press
Page 205 of the Agilent Technologies 8753D user manual — RETURN
Page 205 · click to enlarge
Page 205

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SA VE/RECALL

SELECT DISK .

2. Choose from the following storage devices:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INTERNAL MEMORY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INTERNAL DISK

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EXTERNAL DISK

N

RETURN and then use the

NNNNNNNNNNNNNNNNNNN

* keys or the front panel knob to high-light the name

of the le that you want to rename .

NNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RETURN

FILE UTILITIES

RENAME FILE

ERASE TITLE .

5. Turn the front panel knob to point to each character of the new lename , pressing

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SELECT LETTER when the arrow points to each character . Press

BACK SPACE if you enter

an incorrect character . After you have selected all the characters in the new lename , press

NNNNNNNNNNNNNN

DONE .

Renaming les may also be done by using the optional external keyboard.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SA VE/RECALL

SELECT DISK .

2. Choose from the following storage devices:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INTERNAL MEMORY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INTERNAL DISK

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EXTERNAL DISK

3. Press the

* keys or the front panel knob to high-light the name of the le that you

want to recall.

NNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

4. Press

RETURN

RECALL STATE .

4-42 Printing, Plotting, and Saving Measurement Results

Show 12 more labels / short text on this page
Renaming a File · Press · · · Press · + · Press · Recalling a File · Press · · + ·
Page 206 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 206 · click to enlarge
Page 206

F ormatting a Disk

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SA VE/RECALL

FILE UTILITIES

FORMAT DISK .

2. Choose the type of format you want:

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FORMAT:LIF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FORMAT:DOS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNN

FORMAT EXT DISK

YES .

Solving Problems with Saving or Recalling Files

If you encounter a problem when you are storing les to disk, or the analyzer internal memory ,

check the following list for possible causes:

Look in the analyzer display message area. The analyzer may show a message that will

identify the problem. Refer to the \Error Messages" chapter if you view a message .

Make sure that you are NOT using a single-sided oppy disk in the analyzer disk drive .

Make sure that you are using a formatted disk.

Make sure that the disk has not been formatted with the LIF-1 (hierarchical le system)

extensions as the analyzer does not support this format.

If Y ou are Using an External Disk Drive

Make sure that the analyzer is in system controller mode , by pressing

LOCAL

N

SYSTEM CONTROLLER .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Make sure that you have connected the disk drive to ac power , switched on the power , and

connected an HP-IB cable between the disk drive and the analyzer .

Make sure that the analyzer recognizes the disk drive's HP-IB address , as explained earlier in

this chapter .

Make sure that the analyzer recognizes the disk (drive) unit that you selected (0 or 1).

If the external disk is a hard disk, make sure that the disk volume number is set correctly .

If the disk drive is an older HP 9122, it may not recognize the newer high density disks .

Substitute the HP-IB cable .

Substitute the disk drive .

Printing, Plotting, and Saving Measurement Results 4-43

Show 4 more labels / short text on this page
Press · · Press ·
Page 207 of the Agilent Technologies 8753D user manual — Chapter 4 Printing, Plotting, and Saving Measurement Results
Page 207 · click to enlarge
Page 207
Page 207 of the Agilent Technologies 8753D user manual — Chapter 4 Printing, Plotting, and Saving Measurement Results
Page 208 of the Agilent Technologies 8753D user manual — Chapter 5 Optimizing Measurement Results
Page 208 · click to enlarge
Page 208

This chapter describes techniques and analyzer functions that help you achieve the best

measurement results . The following topics are included in this chapter:

P erformance veri cation

Reference plane and port extensions

Frequency response and isolation correction

One-port re ection correction

TRL* and TRM* error-correction

Modifying calibration kit standards

P ower meter measurement calibration

Calibrating for noninsertable devices

Modify the cal kit thru de nition

Maintaining testport output power during sweep retrace

Making accurate measurements of electrically long devices

Optimizing Measurement Results 5-1

Show 16 more labels / short text on this page
Optimizing Measurement Results · Increasing measurement accuracy · Connector repeatability · Interconnecting cables · T emperature drift · Frequency drift · Measurement error-correction · Frequency response correction · Full two-port correction · A dapter removal · Matched adapters · Increasing sweep speed · Increasing dynamic range · Reducing trace noise · Reducing receiver crosstalk · Reducing recall time
Page 209 of the Agilent Technologies 8753D user manual — Chapter 5 Optimizing Measurement Results
Page 209 · click to enlarge
Page 209

Where to Look for More Information

A dditional information about many of the topics discussed in this chapter is located in the

following areas:

Chapter 2, \Making Measurements ," contains step-by-step procedures for making

measurements or using particular functions .

Chapter 4, \Printing, Plotting, and Saving Measurement Results ," contains instructions

for saving to disk or to the analyzer internal memory , and printing and plotting displayed

measurements .

Chapter 6, \Application and Operation Concepts ," contains explanatory-style information

about many applications and analyzer operation.

Increasing Measurement A ccuracy

Connector repeatability is a source of random measurement error . Measurement

error-corrections do not compensate for these errors . F or all connectors , you should frequently

do the following:

use correct connection techniques (See Chapter 2, T able 2-1.)

Cables connecting the device under test to the analyzer can contribute random errors to your

measurement. Y ou should frequently do the following:

inspect for lossy cables

inspect for damaged cable connectors

practice good connector care techniques

minimize cable position changes between error-correction and measurements

inspect for cables which dramatically change magnitude or phase response when exing (This

may indicate an intermittent problem.)

Electrical characteristics will change with temperature due to the thermal expansion

characteristics of devices within the analyzer , calibration devices , test devices , cables , and

adapters . Therefore , the operating temperature is a critical factor in their performance . During

a measurement calibration, the temperature of the calibration devices must be stable and

within 25 5

use a temperature-controlled environment

ensure the temperature stability of the calibration devices

avoid handling the calibration devices unnecessarily during calibration

ensure the ambient temperature is 1

C of measurement error-correction temperature

5-2 Optimizing Measurement Results

Show 8 more labels / short text on this page
Connector Repeatability · inspect the connectors · clean the connectors · gauge the connectors · Interconnecting Cables · T emperature Drift · ·
Page 210 of the Agilent Technologies 8753D user manual — CAL
Page 210 · click to enlarge
Page 210

Minute changes in frequency accuracy and stability can occur as a result of temperature and

aging (on the order of parts per million). If you require greater frequency accuracy , do the

following:

Override the internal crystal with a high-stability external source , frequency standard, or (if

your analyzer is equipped with Option 1D5) use the internal frequency standard.

P erformance V eri cation

Y ou should periodically check the accuracy of the analyzer measurements , by doing the

following:

perform a measurement veri cation at least once per year

Refer to the HP 8753D Service Guide for the measurement veri cation procedure .

Reference Plane and P ort Extensions

Use the port extension feature to compensate for the phase shift of an extended measurement

reference plane , due to such additions as cables , adapters , and xtures , after completing an

error-correction procedure (or when there is no active correction).

Using port extensions is similar to using electrical delay . However , using port extensions is the

preferred method of compensating for test xture phase shift. T able 5-1 explains the di erence

between port extensions and electrical delay .

T able 5-1.

Di erences between PORT EXTENSIONS and ELECTRICAL DELA Y

FFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFF

PORT EXTENSIONS

F

FFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFF

ELECTRICAL DELAY

Main E ect The end of a cable becomes the test port plane

Compensates for the electrical length of a

for all S-parameter measurements .

cable .

Set the cable's electrical length x 1 for

transmission.

Set the cable's electrical length x 2 for

re ection.

All S-parameters . Only the currently selected S-parameter .

Intelligently compensates for 1 times or 2

Only compensates for electrical length.

times the cable's electrical delay , depending

on which S-parameter is computed.

NNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Y ou can activate a port extension by pressing

CAL

MORE

PORT EXTENSIONS

EXTENSIONS ON .

Then enter the delay to the reference plane .

Optimizing Measurement Results 5-3

Show 6 more labels / short text on this page
Frequency Drift · Measurements · A ected · Electrical · Compensation ·
Page 211 of the Agilent Technologies 8753D user manual — Chapter 5 Optimizing Measurement Results
Page 211 · click to enlarge
Page 211

The accuracy of network analysis is greatly in uenced by factors external to the network

analyzer . Components of the measurement setup , such as interconnecting cables and adapters ,

introduce variations in magnitude and phase that can mask the actual response of the device

under test.

Error-correction is an accuracy enhancement procedure that removes systematic errors

(repeatable measurement variations) in the test setup . The analyzer measures known standard

devices , and uses the results of these measurements to characterize the system.

Conditions Where Error-Correction is Suggested

Measurement accuracy and system characteristics can be a ected by the following factors:

A dapting to a di erent connector type or impedance .

Connecting a cable between the test device and an analyzer test port.

Connecting any attenuator or other such device on the input or output of the test device .

If your test setup meets any of the conditions above , the following system characteristics may

be a ected:

amplitude at device input

Several types of error correction are available that remove from one to twelve systematic

errors . The full 2-port correction e ectively removes all twelve correctable systematic errors .

Some measurements do not require correction for all twelve errors . The following table

explains each correction and its uses .

5-4 Optimizing Measurement Results

Show 7 more labels / short text on this page
Measurement Error-Correction · frequency response accuracy · directivity · crosstalk (isolation) · source match · load match · Types of Error-Correction
Page 212 of the Agilent Technologies 8753D user manual — TRL
Page 212 · click to enlarge
Page 212

T able 5-2. Purpose and Use of Di erent Error-Correction Procedures

Response Transmission or re ection

Frequency response . Thru for transmission, open

measurement when the highest

or short for re ection.

accuracy is not required.

Response & isolation Transmission of high insertion loss

Same as response plus

devices or re ection of high return

isolation in transmission or

isolation standard. (load)

loss devices . Not as accurate as

directivity in re ection.

1-port or 2-port correction.

S

1-port Re ection of any one-port device

Directivity , source match,

Short and open and load.

or well terminated two-port

frequency response .

device .

S

1-port Re ection of any one-port device

Directivity , source match,

Short and open and load.

or well terminated two-port

frequency response .

device .

Full 2-port Transmission or re ection of

Directivity , source match,

Short and open and load and

highest accuracy for two-port

thru. (2 loads for isolation)

devices .

forward and reverse .

TRL

/LRM

Transmission or re ection when

Thru, re ect, line , or line ,

highest accuracy is not required.

frequency response .

re ect, match, or thru,

re ect, match.

Frequency response calibration is not as accurate as other calibration methods .

Error-correction is only valid for a speci c stimulus state , which you must select before you

start a correction. If you change any of the following parameters , you will invalidate the

correction and the analyzer will switch the correction o (unless the interpolated error

correction feature is activated):

The error-correction quality may be degraded (Cor changes to C? ), if you change the following

stimulus state parameters:

If you activate averaging, the analyzer may increase the sweep time if more

time is needed to calculate the averages . If the sweep time changes , you will

see Cor change to C? . The number of averages does not a ect a sweep cycle

time . Therefore , if you use averaging for error-correction, leave it on for the

measurement and set the averaging factor to 1, for a faster sweep .

Optimizing Measurement Results 5-5

Show 20 more labels / short text on this page
Correction · Corresponding · Errors Corrected Standard · Procedure · Measurement · Devices · Frequency response plus · load match, isolation, · frequency response , · * · * · Directivity , isolation, · (forward and reverse) · Error-Correction Stimulus State · frequency range · number of points · sweep type · sweep time · system bandwidth · output power
Page 213 of the Agilent Technologies 8753D user manual — CAL
Page 213 · click to enlarge
Page 213

The quality of the error-correction is limited by two factors: (1) the di erence between the

model of the calibration standards and the actual electrical characteristics of those standards ,

and (2) the condition of the calibration standards . T o make the highest quality measurement

calibration, follow the suggestions below:

use the correct standard model

inspect the calibration standards

clean the calibration standards

gauge the calibration standards

use correct connection techniques

If you want to use calibration standards other than the default sets , you must change the

standard model. (Refer to \Modifying Calibration Kit Standards" located later in this chapter .)

After you enter the mathematical model for the new calibration standards , the analyzer can

then use the model that corresponds to the new standards .

Compensating for the Electrical Delay of Calibration Standards

Short and open calibration standards in the 3.5 mm, 2.4 mm, and 2.92 mm connector types

have a certain amount of electrical delay . The analyzer compensates for this delay by o setting

the calibration results by the total amount of electrical delay caused by the calibration standard

in both the forward and reverse direction. As a result, if these standards are measured after a

calibration, they will not appear to be \perfect" shorts or opens . This is an indication that your

analyzer is working properly and that it has successfully performed a calibration.

If you enter the opposite amount of electrical delay that was used by the

analyzer during calibration, then the short calibration standard will appear

to be \perfect." The open calibration standard has additional phase shift

caused by fringing capacitance . See \Calibration Considerations" in Chapter 6,

\Application and Operation Concepts ."

Clarifying Type-N Connector Sex

When you are performing error-correction for a system that has type-N test port connectors ,

the softkey menus label the sex of the test port connector - not the calibration standard

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

connector . F or example , the label

SHORT (F) refers to the short that will be connected to the

female test port.

When to Use Interpolated Error-Correction

Y ou may want to use interpolated error-correction when you choose a subset of a frequency

range that you already corrected, when you change the number of points , or when you change

to CW . This feature also allows you to change the parameters in a 2-port correction, such as IF

bandwidth, power , or sweep time . The analyzer calculates the systematic errors from the errors

of the original correction.

The quality of the interpolated error-correction depends on the amount of phase shift and

amplitude change of the error coecients between measurement points . If the phase shift

per ve measurement points , the interpolated error-correction can be a great

improvement over uncorrected measurement.

T o activate interpolated measurement correction, press

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INTERPOL ON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CORRECTION ON .

CAL

When interpolation is in use , the notation C? will appear on the analyzer display .

5-6 Optimizing Measurement Results

Show 4 more labels / short text on this page
Calibration Standards · is <180 · ·
Page 214 of the Agilent Technologies 8753D user manual — Chapter 5 Optimizing Measurement Results
Page 214 · click to enlarge
Page 214

Procedures for Error-Correcting Y our Measurements

This section has example procedures or information on the following topics:

frequency response and isolation correction

one-port re ection correction

modifying calibration kit standards

power meter measurement calibration procedure

If you are making measurements on uncoupled measurement channels , you

must make a correction for each channel.

Optimizing Measurement Results 5-7

frequency response correction · full two-port correction · TRL*/LRM* correction

Page 215 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 215 · click to enlarge
Page 215

Y ou can remove the frequency response of the test setup for the following measurements .

combined re ection and transmission measurements

Response Error-Correction for Re ection Measurements

1. Press

.

PRESET

2. Select the type of measurement you want to make .

If you want to make a re ection measurement on PORT 1 (in the forward direction, S

leave the instrument default setting.

If you want to make a re ection measurement on PORT 2 (in the reverse direction S

press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ReflREV S22 (B/R)

MEAS

Set any other measurement parameters that you want for the device measurement:

power , sweep type , number of points , or IF bandwidth.

T o access the measurement error-correction menus , press:

CAL

If your calibration kit is di erent than the kit speci ed under the

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL KIT [ ] softkey ,

press:

NNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL KIT

SELECT CAL KIT (select your type of kit)

NNNNNNNNNNNNNNNNNNNN

RETURN

If your type of calibration kit is not listed in the displayed menu, refer to the \Modifying

Calibration Standards" procedure , located later in this chapter .

T o select a response correction, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNN

CAL

CALIBRATE MENU

RESPONSE

Connect the short or open calibration standard to the port you selected for the test port

(PORT 1 for S

or PORT 2 for S

).

Include any adapters or cables that you will have in the device measurement.

That is , connect the standard device to the particular connector where you will

connect your device under test.

5-8 Optimizing Measurement Results

Show 11 more labels / short text on this page
Frequency Response Error-Corrections · re ection measurements · transmission measurements · ), · ), · · · · · ·
Page 216 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNN
Page 216 · click to enlarge
Page 216
Agilent Technologies 8753D — Chapter 5 Optimizing Measurement Results diagram (page 216)
Fig. 102 · Chapter 5 Optimizing Measurement Results

Figure 5-1.

Standard Connections for a Response Error-Correction for Re ection Measurement

T o measure the standard when the displayed trace has settled, press:

NNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNN

OPEN

If the calibration kit you selected has a choice between male and female calibration

standards , remember to select the sex that applies to the test port and not the standard.

The analyzer displays WAIT - MEASURING CAL STANDARD during the standard measurement.

The analyzer underlines the softkey that you selected after it nishes the measurement,

and computes the error coecients .

This calibration allows only one standard to be measured. If you press the

wrong key for a standard, start over with step 6. Do not use a thru standard

for a re ection response correction.

Y ou can save or store the measurement correction to use for later

measurements , that use the same measurement parameters . Refer to the

\Printing, Plotting, and Saving Measurement Results" chapter for procedures .

This completes the response correction for re ection measurements . Y ou can connect and

measure your device under test.

Optimizing Measurement Results 5-9

SHORT or

Page 217 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNN
Page 217 · click to enlarge
Page 217

Response Error-Correction for Transmission Measurements

1. Press

PRESET

.

2. Select the type of measurement you want to make .

If you want to make a transmission measurement in the forward direction (S

), press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TransFWD S21 (B/R)

MEAS

If you want to make a transmission measurement in the reverse direction (S

), press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MEAS

TransREV S12 (A/R)
3. Set any other measurement parameters that you want for the device measurement: power ,

number of points , IF bandwidth.

4. T o select a response correction, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNN

CAL

CALIBRATE MENU

RESPONSE

5. Make a \thru" connection between the points where you will connect your device under

test.

Include any adapters or cables that you will have in the device measurement.

That is , connect the standard device where you will connect your device under

test.

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 217)
Fig. 103

Figure 5-2.

Standard Connections for Response Error-Correction for Transmission Measurements

6. T o measure the standard, press:

NNNNNNNNNNNNNN

THRU

The analyzer displays WAIT - MEASURING CAL STANDARD during the standard measurement.

NNNNNNNNNNNNNN

The analyzer underlines the

THRU softkey after it measures the calibration standard, and

computes the error coecients .

5-10 Optimizing Measurement Results

Show 6 more labels / short text on this page
· · · · ·
Page 218 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 218 · click to enlarge
Page 218

Do not use an open or short standard for a transmission response correction.

Y ou can save or store the measurement correction to use for later

measurements . Refer to the \Printing, Plotting, and Saving Measurement

Results" chapter for procedures .

7. This completes the response correction for transmission measurements . Y ou can connect and

measure your device under test.

Receiver calibration provides a frequency response error-correction that also indicates absolute

power in dBm. This calibration is most useful when performed with a power meter calibration.

1. P erform a power meter calibration. See \P ower Meter Measurement Calibration," located

later in this chapter .

2. T o set the analyzer test port power to 0 dBm, press:

NNNNNNNNNNNNNNNNN

POWER

MENU

3. Make a \thru" connection between the points where you will connect your device under

test.

Include any adapters or cables that you will have in the device measurement.

That is , connect the standard device where you will connect your device under

test.

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNN diagram (page 218)
Fig. 104

Figure 5-3. Standard Connections for Receiver Calibration

4. T o choose a non-ratioed measurement, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INPUT PORTS

NNNNN

B

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TEST PORT 1

MEAS

This sets the source at PORT 1, and the measurement receiver to PORT 2, or B channel.

5. Set any other measurement parameters that you want for the device measurement: power ,

number of points , IF bandwidth.

6. T o perform a receiver error-correction, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL

RECEIVER CAL

TAKE RCVR CAL SWEEP

Optimizing Measurement Results 5-11

Show 11 more labels / short text on this page
Receiver Calibration · · · 0 · · · · · · 0 ·
Page 219 of the Agilent Technologies 8753D user manual — Chapter 5 Optimizing Measurement Results
Page 219 · click to enlarge
Page 219

Y ou can save or store the measurement correction to use for later

measurements . Refer to the \Printing, Plotting, and Saving Measurement

Results" chapter for procedures .

7. This completes the receiver calibration for transmission measurements . Y ou can connect and

measure your device under test.

The accuracy of the receiver calibration will be nearly the same as the test

port power accuracy; and the test port power accuracy can be signi cantly

improved by performing a power meter source calibration, as described later in

5-12 Optimizing Measurement Results

this chapter

Page 220 of the Agilent Technologies 8753D user manual — N
Page 220 · click to enlarge
Page 220

Frequency Response and Isolation Error-Corrections

removes frequency response of the test setup

removes isolation in transmission measurements

removes directivity in re ection measurements

Y ou can make a response and isolation correction for the following measurements .

combined re ection and transmission measurements

Response and Isolation Error-Correction for Re ection Measurements

Although you can perform a response and isolation correction for re ection measurements ,

Hewlett-P ackard recommends that you perform an S

one-port error-correction; it is more

accurate and just as convenient.

1. Press

.

PRESET

2. Select the type of measurement you want to make .

If you want to make a re ection measurement on PORT 1 (in the forward direction, S

leave the instrument default setting.

If you want to make a re ection measurement on PORT 2 (in the reverse direction, S

press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MEAS

ReflREV S22 (B/R)
3. Set any other measurement parameters that you want for the device measurement: power ,

sweep type , number of points , IF bandwidth.

4. T o access the measurement correction menus , press:

CAL

If your calibration kit is di erent than the kit speci ed under the

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL KIT [ ] softkey ,

press:

NNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNN

CAL KIT

SELECT CAL KIT (select your type of kit)

RETURN

If your type of calibration kit is not listed in the displayed menu, refer to the \Modifying

Calibration Kit Standards" procedure , located later in this chapter .

6. T o select a response and isolation correction, press:

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNN

CALIBRATE MENU

RESPONSE & ISOL'N

RESPONSE

7. Connect the short or open calibration standard to the port you selected for the test port

(PORT 1 for S

or PORT 2 for S

).

Include any adapters that you will have in the device measurement. That is ,

connect the standard device to the particular connector where you will connect

your device under test.

Optimizing Measurement Results 5-13

Show 8 more labels / short text on this page
re ection measurements · transmission measurements · ), · ), · · · ·
Page 221 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 221 · click to enlarge
Page 221
Agilent Technologies 8753D — Chapter 5 Optimizing Measurement Results diagram (page 221)
Fig. 105 · Chapter 5 Optimizing Measurement Results

Figure 5-4.

Standard Connections for a Response and Isolation Error-Correction for Re ection

8. T o measure the standard, press:

NNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNN

OPEN

If the calibration kit you selected has a choice between male and female calibration

standards , remember to select the sex that applies to the test port and not the standard.

The analyzer displays WAIT - MEASURING CAL STANDARD during the standard measurement.

The analyzer underlines the softkey that you selected after it nishes the measurement,

and computes the error coecients .

9. Connect the load calibration standard to the test port.
10. T o measure the standard for the isolation portion of the correction, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ISOL'N STD

11. T o compute the response and directivity error coecients , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DONE RESP ISOL'N CAL

The analyzer displays the corrected S

) data. The analyzer also shows the notation

Cor to the left of the screen, indicating that the correction is switched on for this channel.

Y ou can save or store the error-correction to use for later measurements .

Refer to the \Printing, Plotting, and Saving Measurement Results" chapter for

procedures .

12. This completes the response and isolation error-correction for re ection measurements . Y ou

can connect and measure your device under test.

5-14 Optimizing Measurement Results

Measurements · SHORT or · (or S

Page 222 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNN
Page 222 · click to enlarge
Page 222

Response and Isolation Error-Correction for Transmission Measurements

This procedure is intended for measurements that have a measurement range of greater than

90 dB .

1. Press

PRESET

.

2. Select the type of measurement you want to make .

If you want to make a transmission measurement in the forward direction (S

), press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MEAS

TransFWD S21 (B/R)

If you want to make a transmission measurement in the reverse direction (S

), press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MEAS

TransREV S12 (A/R)
3. Set any other measurement parameters that you want for the device measurement: power ,

number of points , IF bandwidth.

4. T o access the measurement correction menus , press:

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

If your calibration kit is di erent than the kit speci ed under the

CAL KIT [ ] softkey ,

press:

NNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL KIT

SELECT CAL KIT (select your type of kit)

If your type of calibration kit is not listed in the displayed menu, refer to the \Modifying

Calibration Kit Standards" procedure , located later in this chapter .

6. T o select a response and isolation correction, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CALIBRATE MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RESPONSE & ISOL'N

NNNNNNNNNNNNNNNNNNNNNNNNNN

RESPONSE

7. Make a \thru" connection between the points where you will connect your device under

test.

Include any adapters that you will have in the device measurement. That is ,

connect the standard device to the particular connector where you will connect

your device under test.

8. T o measure the standard, when the displayed trace has settled, press:

NNNNNNNNNNNNNN

THRU

The analyzer displays WAIT - MEASURING CAL STANDARD during the standard measurement.

NNNNNNNNNNNNNN

The analyzer underlines the

THRU softkey after it measures the calibration standard, and

computes the error coecients .

9. Connect impedance-matched loads to PORT 1 and PORT 2, as shown in Figure 5-5. Include

the adapters that you would include for your device measurement.

Optimizing Measurement Results 5-15

Show 6 more labels / short text on this page
· · · · ·
Page 223 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 223 · click to enlarge
Page 223
Agilent Technologies 8753D — Chapter 5 Optimizing Measurement Results diagram (page 223)
Fig. 106 · Chapter 5 Optimizing Measurement Results

Figure 5-5.

Standard Connections for a Response and Isolation Error-Correction for Transmission

If you will be measuring highly re ective devices , such as lters , use the test

device , connected to the reference plane and terminated with a load, for the

isolation standard.

10. T o help remove crosstalk noise , set the analyzer as follows:

a.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

A VG

AVERAGING ON

AVERAGING FACTOR and enter at least four times more

averages than desired during the device measurement.

NNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

b.

CAL

MORE

ALTERNATE A and B to eliminate one crosstalk path.

11. T o measure the calibration standard, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL

RESUME CAL SEQUENCE

ISOL'N STD

12. Return the averaging to the original state of the measurement. F or example , reduce the

averaging factor by at least four times or turn averaging o .

13. T o compute the isolation error coecients , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL

RESUME CAL SEQUENCE

DONE RESP ISOL'N CAL

The analyzer displays the corrected data trace . The analyzer also shows the notation Cor at

the left of the screen, indicating that the correction is switched on for this channel.

Y ou can save or store the measurement correction to use for later

measurements . Refer to the \Printing, Plotting, and Saving Measurement

Results" chapter for procedures .

14. This completes the response and isolation correction for transmission measurements . Y ou

can connect and measure your device under test.

5-16 Optimizing Measurement Results

Show 9 more labels / short text on this page
Measurements · Press · · Press · · · · ·
Page 224 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 224 · click to enlarge
Page 224

One-P ort Re ection Error-Correction

removes directivity errors of the test setup

removes source match errors of the test setup

removes frequency response of the test setup

Y ou can perform a 1-port correction for either an S

measurement. The only

di erence between the two procedures is the measurement parameter that you select.

This is the recommended error-correction process for all re ection

measurements , when full two-port correction is not used.

1. Press

PRESET

.

2. Select the type of measurement you want to make .

If you want to make a re ection measurement on PORT 1 (in the forward direction, S

leave the instrument default setting.

If you want to make a re ection measurement on PORT 2 (in the reverse direction, S

press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ReflREV S22 (B/R)

MEAS

3. Set any other measurement parameters that you want for the device measurement: power ,

number of points , IF bandwidth.

4. T o access the measurement correction menus , press:

CAL

If your calibration kit is di erent than the kit speci ed under the

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL KIT [ ] softkey ,

press:

NNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SELECT CAL KIT (select your type of kit)

NNNNNNNNNNNNNNNNNNNN

CAL KIT

RETURN

If your type of calibration kit is not listed in the displayed menu, refer to the \Modifying

Calibration Kit Standards" procedure , located later in this chapter .

6. T o select the correction type , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CALIBRATE MENU and select the correction type

If you want to make a re ection measurement at PORT 1 press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

S11 1-PORT

If you want to make a re ection measurement at PORT 2, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

S22 1-PORT

7. Connect a shielded open circuit to PORT 1 (or PORT 2 for an S

measurement).

Include any adapters that you will have in the device measurement. That is ,

connect the calibration standard to the particular connector where you will

connect your device under test.

Optimizing Measurement Results 5-17

Show 7 more labels / short text on this page
or an S · ), · ), · · · ·
Page 225 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 225 · click to enlarge
Page 225
Agilent Technologies 8753D — Chapter 5 Optimizing Measurement Results diagram (page 225)
Fig. 107 · Chapter 5 Optimizing Measurement Results

Figure 5-6. Standard Connections for a One P ort Re ection Error-Correction

8. T o measure the standard, when the displayed trace has settled, press:

NNNNNNNNNNNNN

N

OPEN

If the calibration kit that you selected has a choice between male or female

calibration standards , remember to select the sex that applies to the test port

and not the standard.

The analyzer displays WAIT - MEASURING CAL STANDARD during the standard measurement.

The analyzer underlines the

NNNNNNNNNNNNNN

OPEN softkey after it measures the calibration standard.

9. Disconnect the open, and connect a short circuit to the test port.
10. T o measure the standard when the displayed trace has settled, press:

N

NNNNNNNNNNNNNNNN

SHORT

NNNNNNNNNNNNNNNNN

The analyzer measures the short circuit and underlines the

SHORT softkey .

11. Disconnect the short, and connect an impedance-matched load to the test port.

When the displayed trace settles , press

NNNNNNNNNNNNNN

LOAD .

NNNNNNNNNNNNNN

The analyzer measures the load and underlines the

LOAD softkey .

13. T o compute the error coecients , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DONE: 1-PORT CAL

The analyzer displays the corrected data trace . The analyzer also shows the notation Cor to

the left of the screen, indicating that the correction is switched on for this channel.

The open, short, and load could be measured in any order , and need not follow

the order in this example .

5-18 Optimizing Measurement Results

Page 226 of the Agilent Technologies 8753D user manual — Chapter 5 Optimizing Measurement Results
Page 226 · click to enlarge
Page 226

Y ou can save or store the error-correction to use for later measurements .

Refer to the \Printing, Plotting, and Saving Measurement Results" chapter for

procedures .

14. This completes the one-port correction for re ection measurements . Y ou can connect and

measure your device under test.

Optimizing Measurement Results 5-19

Page 227 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 227 · click to enlarge
Page 227

Full Two-P ort Error-Correction

removes directivity errors of the test setup in forward and reverse directions

removes source match errors of the test setup in forward and reverse directions

removes load match errors of the test setup in forward and reverse directions

removes isolation errors of the test setup in forward and reverse directions (optional)

removes frequency response of the test setup in forward and reverse directions

This is the most accurate error-correction procedure . Since the analyzer takes

both forward and reverse sweeps , this procedure takes more time than the

other correction procedures .

1. Set any measurement parameters that you want for the device measurement: power ,

format, number of points , IF bandwidth.

2. T o access the measurement correction menus , press:

CAL

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

If your calibration kit is di erent than the kit speci ed under the

CAL KIT [ ] softkey ,

press:

NNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNN

CAL KIT

SELECT CAL KIT (select your type of kit)

RETURN

If your type of calibration kit is not listed in the displayed menu, refer to the \Modifying

Calibration Kit Standards" procedure , located later in this chapter .

4. T o select the correction type , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CALIBRATE MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FULL 2-PORT

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REFLECTION

5. Connect a shielded open circuit to PORT 1.

Include any adapters that you will have in the device measurement. That is ,

connect the standard to the particular connector where you will connect your

device under test.

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 227)
Fig. 108

Figure 5-7. Standard Connections for Full Two port Error-Correction

5-20 Optimizing Measurement Results

·

Page 228 of the Agilent Technologies 8753D user manual — TRANSMISSION
Page 228 · click to enlarge
Page 228
6. T o measure the standard, when the displayed trace has settled, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FORWARD: OPEN

The analyzer displays WAIT - MEASURING CAL STANDARD during the standard measurement.

NNNNNNNNNNNNNN

The analyzer underlines the

OPEN softkey after it measures the standard.

7. Disconnect the open, and connect a short circuit to PORT 1.
8. T o measure the device , when the displayed trace has settled, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FORWARD: SHORT

The analyzer measures the short circuit and underlines the

N

SHORT softkey .

NNNNNNNNNNNNNNNN

9. Disconnect the short, and connect an impedance-matched load to PORT 1.
10. T o measure the standard, when the displayed trace has settled, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FORWARD: LOAD

NNNNNNNNNNNNNN

The analyzer measures the load and underlines the

LOAD softkey .

11. Repeat the open-short-load measurements described above , but connect the devices in turn

to PORT 2, and use the

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REVERSE: OPEN ,

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REVERSESHORT , and

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REVERSE: LOAD softkeys .

Include any adapters that you would include in your device measurement.

12. T o compute the re ection correction coecients , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STANDARDS DONE

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

T o start the transmission portion of the correction, press:

TRANSMISSION

14. Make a \thru" connection between the points where you will connect your device under

test as shown in Figure 5-7.

Include any adapters or cables that you will have in the device measurement.

That is , connect the standard device where you will connect your device under

test.

Optimizing Measurement Results 5-21

Page 229 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 229 · click to enlarge
Page 229

The thru in most calibration kits is de ned with zero length. The correction

will not work properly if a non-zero length thru is used, unless the calibration

kit is modi ed to change the de ned thru to the length used. This is important

for measurements of noninsertable devices (devices having ports that are both

male or both female). The modi ed calibration kit must be saved as the user

NNNNNNNNNNNNNNNNNNNNNNNNNN

calibration kit, and the

USER KIT must be selected before the calibration is

started.

15. T o measure the standard, when the trace has settled, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DO BOTH FWD+REV

The analyzer underlines the softkey label after it makes each measurement.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ISOLATION and select from the following two options:

If you will be measuring devices with a dynamic range less than 90 dB , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

OMIT ISOLATION

If you will be measuring devices with a dynamic range greater than 90 dB , follow these

steps:

a. Connect impedance-matched loads to PORT 1 and PORT 2. Include the adapters that

you would include for your device measurement.

If you will be measuring highly re ective devices , such as lters , use the test

device , connected to the reference plane and terminated with a load, for the

isolation standard.

b. A ctivate at least four times more averages than desired during the device

measurement.

If loads can be connected to both port 1 and port 2 simultaneously , then the

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

following step can be performed using the

DO BOTH FWD + REV softkey .

c. Press

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RESUME CAL SEQUENCE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ISOLATION

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FWD ISOL'N ISOL'N STD

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REV ISOL'N ISOL'N STD

ISOLATION DONE .

d. Return the averaging to the original state of the measurement, and press

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RESUME CAL SEQUENCE .

17. T o compute the error coecients , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DONE 2-PORT CAL

The analyzer displays the corrected measurement trace . The analyzer also shows the

Y ou can save or store the measurement correction to use for later

measurements . Refer to the \Printing, Plotting, and Saving Measurement

Results" chapter for procedures .

18. This completes the full two-port correction procedure . Y ou can connect and measure your

device under test.

5-22 Optimizing Measurement Results

Press · ·

Page 230 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 230 · click to enlarge
Page 230

TRL* and TRM* Error-Correction

The HP 8753D analyzer has the capability of making calibrations using the TRL*/LRM* method.

NNNNNNNNNNNNNNNNNNNNNNNNNN

Y ou must have a TRL calibration kit de ned and saved in the

USER KIT , as shown in

\Modifying Calibration Kit Standards ," located later in this section.

N

NNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL

CAL KIT

SELECT CAL KIT

USER KIT

RETURN

RETURN

CALIBRATE MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRL*/LRM* 2-PORT .

3. T o measure the \TRL THRU ," connect the \zero length" transmission line between the two

test ports .

4. T o make the necessary four measurements , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

THRU TRLTHRU

5. T o measure the \TRL SHORT ," connect the short to PORT 1, and press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

S11 REFL: TRLSHORT

6. Connect the short to PORT 2, and press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

S22 REFL: TRLSHORT

7. T o measure the \TRL LINE," disconnect the short and connect the TRL line from PORT 1 to

PORT 2.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

8. Press

LINE/MATCH

DO BOTH FWD+REV .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

The line data is measured and the

labels are underlined.

10. T o measure the \ISOLA TION" class , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ISOLATION

Y ou could choose not to perform the isolation measurement by pressing

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

OMIT ISOLATION

DONE TRL/LRM .

Y ou should perform the isolation measurement when the highest dynamic range

is desired.

T o perform the best isolation measurements , you should reduce the system

bandwidth, and/or activate the averaging function.

A poorly measured isolation class can actually degrade the overall

measurement performance . If you are in doubt of the isolation measurement

quality , you should omit the isolation portion of this procedure .

If loads can be connected to both port 1 and port 2 simultaneously , then the

following measurement can be performed using the

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DO BOTH FWD + REV

softkey .

Optimizing Measurement Results 5-23

Show 5 more labels / short text on this page
TRL Error-Correction · Press · · LN/MATCH1 TRLLINE and · LN/MATCH2 TRLLINE softkey
Page 231 of the Agilent Technologies 8753D user manual — N
Page 231 · click to enlarge
Page 231
11. Connect a load to PORT 2, and press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REV ISOL'N ISOL'N STD

12. Connect the load to PORT 1, and press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FWD ISOL'N ISOL'N STD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ISOLATION DONE

13. Y ou may repeat any of the steps above . There is no requirement to go in the order of

steps . When the analyzer detects that you have made all the necessary measurements , the

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

message line will show PRESS 'DONE' IF FINISHED WITH CAL . Press

DONE TRL/LRM .

The message COMPUTING CAL COEFFICIENTS will appear , indicating that the analyzer is

performing the numerical calculations of error coecients .

Y ou can save or store the measurement correction to use for later

measurements . Refer to the \Printing, Plotting, and Saving Measurement

Results" chapter for procedures .

14. Connect the device under test. The device S-parameters are now being measured.

Y ou must have a TRM calibration kit de ned and saved in the

NNNNNNNNNNNNNNNNNNNNNNNNNN

USER KIT , as shown in

\Modifying Calibration Kit Standards ," located later in this section.

This must be done before performing the following sequence .

2. Press

NNNNNNNNNNNNNNNNNNNNNNN

CAL KIT

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SELECT CAL KIT

NNNNNNNNNNNNNNNNNNNNNNNNNN

USER KIT

NNNNNNNNNNNNNNNNNNNN

RETURN

NNNNNNNNNNNNNNNNNNNN

RETURN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CALIBRATE MENU

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRL*/LRM* 2-PORT .

3. T o measure the \TRM THRU ," connect the \zero length" transmission line between the two

test ports .

4. T o make the necessary four measurements , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

THRU TRMTHRU

5. T o measure the \TRM SHORT ," connect the short to PORT 1, and press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

S11 REFL: TRMSHORT

6. Connect the short to PORT 2, and press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

S22 REFL: TRMSHORT

If loads can be connected to both port 1 and port 2 simultaneously ,

then the following TRM load measurement can be performed using the

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DO BOTH FWD + REV softkey .

7. T o measure the \TRM LO AD ," disconnect the short and connect the TRM load to PORT1.

Refer to \Choosing Calibration Load Standards ."

5-24 Optimizing Measurement Results

TRM Error-Correction ·

Page 232 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 232 · click to enlarge
Page 232

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LINE/MATCH

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LN/MATCH1 TRMLOAD to access the Loads menu. When the displayed

trace settles , press the softkey corresponding to the load used. If a sliding load is

NNNNNNNNNNNNNNNNNNNNNNN

SLIDING to access the Sliding Load menu. P osition the slide and press

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SLIDE IS SET .

9. When all the appropriate load measurements are complete , the load data is measured and

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LN/MATCH1 TRMLOAD softkey label is underlined.

Connect the load to PORT 2 and press

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LN/MATCH2 TRMLOAD .

11. Repeat the previous TRM load measurement steps for PORT 2.
12. After the measurement is complete , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DONE LINE/MATCH

13. T o measure the \ISOLA TION" class , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ISOLATION

Y ou could choose not to perform the isolation measurement by pressing

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

OMIT ISOLATION

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DONE TRL/LRM .

Y ou should perform the isolation measurement when the highest dynamic range

is desired.

T o perform the best isolation measurements , you should reduce the system

bandwidth, and/or activate the averaging function.

A poorly measured isolation class can actually degrade the overall

measurement performance . If you are in doubt of the isolation measurement

quality , you should omit the isolation portion of this procedure .

14. Y ou may repeat any of the steps above . There is no requirement to go in the order of

steps . When the analyzer detects that you have made all the necessary measurements , the

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

message line will show PRESS 'DONE' IF FINISHED WITH CAL. Press

DONE TRL/LRM .

The message COMPUTING CAL COEFFICIENTS will appear , indicating that the analyzer is

performing the numerical calculations of error coecients .

Y ou can save or store the measurement correction to use for later

measurements . Refer to the \Printing, Plotting, and Saving Measurement

Results" chapter for procedures .

15. Connect the device under test. The device S-parameters are now being measured.

When making measurements using the same port with uncoupled channels , the

power level for each channel must fall within the same power range setting of

that single port. An error message will be displayed if you enter two power

levels that do not fall within the same power range .

Optimizing Measurement Results 5-25

Press · used, press · the

Page 233 of the Agilent Technologies 8753D user manual — N
Page 233 · click to enlarge
Page 233

Modifying Calibration Kit Standards

Numerical data for most Hewlett-P ackard calibration kits is provided in the

calibration kit manuals .

The following section provides a summary of the information in the 8510-5A application note ,

as well as HP 8753D menu-speci c information. F or a detailed description of the menus and

softkeys located in this section, as well as information about when user-de ned calibration kits

should be used, refer to Chapter 6, \Application and Operation Concepts ."

The following are de nitions of terms:

A \standard" (represented by a number 1-8) is a speci c , well-de ned, physical device used

to determine systematic errors . F or example , standard 1 is a short in the 3.5 mm calibration

kit.

A standard \type" is one of ve basic types that de ne the form or structure of the model to

be used with that standard (short, open, load, delay/thru, and arbitrary impedance); standard

1 is of the type short in the 3.5 mm calibration kit.

Standard \coecients" are numerical characteristics of the standards used in the model

selected. F or example , the o set delay of the short is 32 ps in the 3.5 mm calibration kit.

A standard \class" is a grouping of one or more standards that determines which of the eight

standards are used at each step of the calibration. F or example , standard number 2 makes up

A re ection class .

Outline of Standard Modi cation

The following steps are used to modify or de ne user kit standard models , contained in the

analyzer memory . It is not possible to alter the built-in calibration kits; all modi cations will be

saved in the user kit.

1. T o modify a cal kit, rst select the prede ned kit to be modi ed. This is not necessary for

de ning a new cal kit.

2. De ne the standards . F or each standard, de ne which \type" of standard it is and its

electrical characteristics .

3. Specify the class where the standard is to be assigned.
4. Store the modi ed cal kit.

N

NNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL

CAL KIT

SELECT CAL KIT .

2. Select the softkey that corresponds to the kit you want to modify .

N

NNNNNNNNNNNNNNNNNNN

RETURN

NNNNNNNNNNNNNNNNNNNN

MODIFY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DEFINE STANDARD .

4. Enter the number of the standard that you want to modify , followed by

. Refer to your

calibration kit manual for the numbers of the speci c standards in your kit. F or example ,

to select a short press

.

5-26 Optimizing Measurement Results

Show 7 more labels / short text on this page
De nitions · the S · Modifying Standards · Press · · Press · 1
Page 234 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 234 · click to enlarge
Page 234

T able 5-3. Typical Calibration Kit Standard and Corresponding Number

5. Press the underlined softkey . F or example , if you selected

in the previous step ,

NNNNNNNNNNNNNNNNN

SHORT should be the underlined softkey .

Do not press a softkey that is not underlined unless you want to change the

\type" of standard.

6. This step applies only to the open . Go to the next step if you selected any other standard.

a.

N

NNNNNNNNNNNNN

OPEN

NNNNNNNN

C0 . Observe the value on the analyzer screen. Use the entry keys on the

analyzer front panel to change the value .

b.

NNNNNNNN

NNNNNNNN

NNNNNNNN

Repeat the modi cation for

C1 ,

7. This step applies only to the load . Go to the next step if you selected any other standard.

Ensure that the correct load type is underlined:

NNNNNNNNNNNNNNNNN

FIXED ,

NNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNN

OFFSET .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SPECIFY OFFSET

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

OFFSET DELAY and observe the value on the analyzer screen. T o

change the value , use the entry keys on the front panel.

9. Repeat the value modi cation for the characteristics listed below:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

OFFSET LOSS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

OFFSET Z0

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MINIMUM FREQUENCY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MAXIMUM FREQUENCY

Ensure that the correct transmission line is underlined:

NNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

WAVEGUIDE .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STD OFFSET DONE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STD DONE (DEFINED) .

12. Repeat steps 4 through 11 for the remaining standards .

Optimizing Measurement Results 5-27

Show 19 more labels / short text on this page
Typical · Default · Standard Type · Standard Number · short (m) 1 · open (m) 2 · broadband load 3 · delay/thru 4 · sliding load 5 · lowband load 6 · short (f) 7 · open (f) 8 · 1 · Press · C2 , and · SLIDING , or · Press · COAX or · Press
Page 235 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNN
Page 235 · click to enlarge
Page 235

Saving the modi ed calibration constants

If you made modi cations to any of the standard de nitions , follow the remaining steps in this

procedure to assign a kit label, and store them in the non-volatile memory . The new set of

NNNNNNNNNNNNNNNNNNNNNNNNNN

standard de nitions will be available under

USER KIT until you save another user kit.

NNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL

CAL KIT

MODIFY

LABEL KIT

ERASE TITLE . Use the front panel knob to

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

move the pointer to a character and press

SELECT LETTER .

T o enter titles , you may also use the optional external keyboard.

NNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

14. Press

DONE

KIT DONE (MODIFIED)

SAVE USER KIT .

Y ou may also save the user kit to disk, by selecting the particular kit at the

time you save a measurement result.

In order to use the TRL technique , the calibration standards characteristics must be entered

into the analyzer's user de ned calibration kit.

This example procedure shows you how to de ne a calibration kit to utilize a set of TRL

(THRU , REFLECT , LINE) standards . This example TRL kit contains the following:

\ ush" short for the REFLECT standard (0 second o set)

50 ohm transmission line with 80 ps of o set delay for the LINE

Hewlett-P ackard strongly recommends that you read product note 8510-8A

before you attempt to modify the standard de nitions . The part number of this

product note is 5091-3645E. Although the product note was written for the HP

8510 family of network analyzers , it also applies to the HP 8753D .

F or a discussion on TRL calibration, refer to \TRL/LRM Calibration" in

Chapter 6, \Application and Operation Concepts ."

Modify the Standard De nitions

1. Press the following keys to start modifying the standard de nitions:

PRESET

NNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL

CAL KIT

MODIFY

DEFINE STANDARD

2. T o select a short, press

. (In this example the REFLECT standard is a SHORT .)

3. Press the following keys:

NNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SHORT

SPECIFY OFFSET

OFFSET DELAY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STD OFFSET DONE

STD DONE (DEFINED)

5-28 Optimizing Measurement Results

Show 12 more labels / short text on this page
Press · · Modifying TRL Standards · zero length THRU · · · · · 1 · · 0 ·
Page 236 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 236 · click to enlarge
Page 236
4. T o de ne the THRU/LINE standard, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DEFINE STANDARD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DELAY/THRU

SPECIFY OFFSET

OFFSET DELAY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STD OFFSET DONE

STD DONE (DEFINED)

5. T o de ne the LINE/MA TCH standard, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DEFINE STANDARD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DELAY/THRU

SPECIFY OFFSET

OFFSET DELAY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STD OFFSET DONE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

F or the purposes of this example , change the name of the standard by pressing

LABEL STD

and modifying the name to \LINE."

7. When the title area shows the new label, press:

N

NNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DONE

STD DONE (DEFINED)

Assign the Standards to the V arious TRL Classes

8. T o assign the calibration standards to the various TRL calibration classes , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNN

NNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SPECIFY CLASS

MORE

MORE

TRL REFLECT

9. Since you previously designated standard #1 for the REFLECT standard, press:
10. Since you previously designated standard #6 for the LINE/MA TCH standard, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRL LINE OR MATCH

11. Since you previously designated standard #4 for the THRU/LINE standard, press:

N

NNNNNNNNNNNNNNNNNNNNNNNNN

TRL THRU

12. T o complete the speci cation of class assignments , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SPECIFY CLASS DONE

Optimizing Measurement Results 5-29

Show 16 more labels / short text on this page
4 · · 0 · · 6 · · · G/n · · · 1 · · 6 · · 4 ·
Page 237 of the Agilent Technologies 8753D user manual — N
Page 237 · click to enlarge
Page 237

T o enter the following label titles , an external keyboard may be used for

convenience .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNN

NNNNNNNNNNNNNN

13. Press

LABELCLASS

MORE

MORE .

14. Change the label of the \TRL REFLECT" class to \TRLSHORT ."
15. Change the label of the \TRL LINE OR MA TCH" class to \TRLLINE."
16. Change the label of the \TRL THRU" class to \TRL THRU ."

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LABEL CLASS DONE .

Label the Calibration Kit

NNNNNNNNNNNNNNNNNNNNNNNNNN

18. Press

LABELKIT and create a label up to 8 characters long. F or this example , enter \TRL

KIT1"

NNNNNNNNNNNNNN

DONE .

19. T o save the newly de ned kit into nonvolatile memory , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

KIT DONE (MODIFIED)

SAVE USER KIT

In order to use the TRL technique , the calibration standards characteristics must be entered

into the analyzer's user de ned calibration kit.

This example procedure shows you how to de ne a calibration kit to utilize a set of TRM

(THRU , REFLECT , MA TCH) standards . This example TRM kit contains the following:

\ ush" short for the REFLECT standard (0 second o set)

50 ohm termination for the MA TCH (in nite length line)

Hewlett-P ackard strongly recommends that you read product note 8510-8A

before you attempt to modify the standard de nitions . The part number of this

product note is 5091-3645E. Although the product note was written for the

HP 8510 family of network analyzers , it also applies to the HP 8753D .

F or a discussion on TRL calibration, refer to \TRL/LRM Calibration" in

Chapter 6, \Application and Operation Concepts ."

Modify the Standard De nitions

1. Press the following keys to start modifying the standard de nitions:

PRESET

N

NNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL

CAL KIT

MODIFY

DEFINE STANDARD

2. T o select a short, press

. (In this example the REFLECT standard is a SHORT .)

3. Press the following keys:

N

NNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SHORT

SPECIFY OFFSET

OFFSET DELAY

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STD OFFSET DONE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STD DONE (DEFINED)

5-30 Optimizing Measurement Results

Show 12 more labels / short text on this page
Label the Classes · Press · Modifying TRM Standards · zero length THRU · · · · · 1 · · 0 ·
Page 238 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 238 · click to enlarge
Page 238
4. T o de ne the THRU/LINE standard, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DEFINE STANDARD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DELAY/THRU

SPECIFY OFFSET

OFFSET DELAY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STD OFFSET DONE

STD DONE (DEFINED)

5. T o de ne the LINE/MA TCH standard, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DEFINE STANDARD

NNNNNNNNNNNNNN

LOAD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

F or the purposes of this example , change the name of the standard by pressing

LABEL STD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ERASE TITLE , if a previous title exists , and then modify the name to \MA TCH".

7. When the title area shows the new label, press:

NNNNNNNNNNNNNN

DONE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STD DONE (DEFINED)

Assign the Standards to the V arious TRM Classes

8. T o assign the calibration standards to the various TRL calibration classes , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNN

NNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SPECIFY CLASS

MORE

MORE

TRL REFLECT

9. Since you previously designated standard #1 for the REFLECT standard, press:
10. Since you previously designated standard #3 for the LINE/MA TCH standard, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRL LINE OR MATCH

11. Since you previously designated standard #4 for the THRU/LINE standard, press:

NNNNNNNNNNNNNNNNNNNNNNNNNN

TRL THRU

12. T o complete the speci cation of class assignments , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SPECIFY CLASS DONE

Optimizing Measurement Results 5-31

Show 13 more labels / short text on this page
4 · · 0 · · 3 · · · 1 · · 3 · · 4 ·
Page 239 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 239 · click to enlarge
Page 239

T o enter the following label titles , an external keyboard may be used for

convenience .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNN

NNNNNNNNNNNNNN

13. Press

LABELCLASS

MORE

MORE .

14. Change the label of the \TRL REFLECT" class to \TRMSHORT ."
15. Change the label of the \TRL LINE OR MA TCH" class to \TRMLO AD ."
16. Change the label of the \TRL THRU" class to \TRMTHRU ."

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LABEL CLASS DONE .

Label the Calibration Kit

NNNNNNNNNNNNNNNNNNNNNNNNNN

18. Press

LABELKIT and create a label up to 8 characters long. F or this example , enter \TRM

KIT1".

NNNNNNNNNNNNNN

DONE

19. T o save the newly de ned kit into nonvolatile memory , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

KIT DONE (MODIFIED)

SAVE USER KIT

5-32 Optimizing Measurement Results

Label the Classes · Press

Page 240 of the Agilent Technologies 8753D user manual — PRESET
Page 240 · click to enlarge
Page 240

P ower Meter Measurement Calibration

Y ou can use the power meter to monitor and correct the analyzer source power to achieve

calibrated absolute power at the test port. Y ou can also use this calibration to set a reference

power for receiver power calibration, and mixer measurement calibration. The power meter

can measure and correct power in two ways:

continuous correction | each sweep mode

sample-and-sweep correction | single sweep mode

The time required to perform a power meter calibration depends on the source power ,

number of points tested, and number of readings taken. The parameters used to derive the

characteristic values in T able 5-4 are as follows:

number of points51, 100 kHz to 3 GHz
test port powerequal to calibration power

T able 5-4.

Characteristic P ower Meter Calibration Sweep Speed and A ccuracy

Number of Readings Sweep Time

at T est P ort (dBm)

1 33 0.7

+5 2 64 0.2

3 95 0.1

1 48 0.7

15 2 92 0.2

3 123 0.1

1 194 0.7

30 2 360 0.2

3 447 0.1

1 Sweep speed applies to every sweep in continuous correction mode , and

to the rst sweep in sample-and-sweep mode . Subsequent sweeps in

sample-and-sweep mode will be much faster .

2 The accuracy values were derived by combining the accuracy of the

power meter and linearity of the analyzer's internal source , as well as

the mismatch uncertainty associated with the power sensor .

Loss of P ower Calibration Data

If your instrument state has not been saved after a power meter calibration,

the power correction data will be lost if any of the following circumstances

exists:

if you switch o the analyzer ac power and you haven't saved the correction

in an internal register

if you press

PRESET

and you haven't saved the correction in an internal

if you change the sweep type (linear , log, list, CW , power) when the power

meter correction is activated

if you change the frequency when the sweep type is in log or list mode

Optimizing Measurement Results 5-33

Show 5 more labels / short text on this page
P ower Desired · Characteristic · (seconds) · A ccuracy (dB) · register
Page 241 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNN
Page 241 · click to enlarge
Page 241

Entering the P ower Sensor Calibration Data

Entering the power sensor calibration data compensates for the frequency response of the

power sensor , thus ensuring the accuracy of power meter calibration.

1. Make sure that your analyzer and power meter are con gured. Refer to the \Compatible

P eripherals" chapter for con guration procedures .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL

PWRMTR CAL

LOSS/SENSR LISTS

CAL FACTOR SENSOR A .

The analyzer shows the notation EMPTY , if you have not entered any segment information.

3. T o create the rst segment, press:

NNNNNNNNNNN

ADD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FREQUENCY

4. Enter the frequency of a correction factor data point, as listed on the power sensor ,
followed by the appropriate key

M/

.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL FACTOR and enter the correction factor that corresponds to the frequency that

you have entered in the previous step . Complete the correction factor entry by pressing

NNNNNNNNNNNNNN

DONE .

6. Repeat the previous two steps to enter up to 55 frequency segments .

Y ou may enter multiple segments in any order because the analyzer automatically sorts

them and lists them on the display by frequency value . The analyzer also automatically

interpolates the values between correction factor data points .

If you only enter one frequency segment the analyzer assumes that the single value is valid

over the entire frequency range of the correction.

NNNNNNNNNNNNNN

After you have entered all the frequency segments , press

DONE .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

A ccess the \Segment Modify Menu" by pressing

CAL

PWRMTR CAL

LOSS/SENSR LISTS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL FACTOR SENSOR A (or

CAL FACTOR SENSOR B , depending on where the segment is

that you want to edit).

NNNNNNNNNNNNNNNNNNNNNNN

Identify the segment that you want to edit by pressing

SEGMENT and using the

* and

keys to locate and position the segment next to the pointer (>), shown on the display . Or

NNNNNNNNNNNNNNNNNNNNNNN

SEGMENT and enter the segment number followed by

.

NNNNNNNNNNNNNN

EDIT and then press either the

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL FACTOR key , depending of which

part of the segment you want to edit.

If you are modifying the frequency , enter the new value , followed by a

M/

key .

If you are modifying the correction factor , enter the new value , followed by the

key .

NNNNNNNNNNNNNN

DONE after you have nished modifying the segment.

NNNNNNNNNNNNNNNNNNNNNNN

If you want to edit any other segments , press

SEGMENT and follow the previous steps ,

starting with step 2.

5-34 Optimizing Measurement Results

Show 21 more labels / short text on this page
Press · · G/n · · · k/m · Press · · Editing Frequency Segments · · + · · press · Press · FREQUENCY or · , · , or · G/n · · k/m · Press
Page 242 of the Agilent Technologies 8753D user manual — CAL
Page 242 · click to enlarge
Page 242

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

A ccess the \Segment Modify Menu" by pressing

CAL

PWRMTR CAL

LOSS/SENSR LISTS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL FACTOR SENSOR A (or

CAL FACTOR SENSOR B , depending on where the segment is

that you want to delete).

Identify the segment that you want to delete by pressing

NNNNNNNNNNNNNNNNNNNNNNN

SEGMENT and using the

* and

keys to locate and position the segment next to the pointer (>), shown on the display . Or

NNNNNNNNNNNNNNNNNNNNNNN

SEGMENT and enter the segment number followed by

.

NNNNNNNNNNNNNNNNNNNN

DELETE .

The analyzer deletes the segment and moves the remainder of the segments up one number .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNN

Y ou could also delete all the segments in a list by pressing

CLEAR LIST

YES .

NNNNNNNNNNNNNN

DONE when you are nished modifying the segment list.

Compensating for Directional Coupler Response

If you use a directional coupler to sample power in your measurement con guration, you

should enter the coupled arm power loss value into the power loss table , using the following

procedure . Y ou can enter the loss information in a single segment, and the analyzer will assume

that the value applies to the entire frequency range of the instrument. Or , you can input actual

measured power loss values at several frequencies using up to 55 segments , enhancing power

accuracy .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL

PWRMTR CAL

LOSS/SENSR LISTS

POWER LOSS .

The analyzer shows the notation EMPTY , if you have not entered any segment information.

NNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

T o create the rst segment, press

ADD

FREQUENCY and enter a frequency of a correction

factor data point, followed by the appropriate key:

.

M/

NNNNNNNNNNNNNN

LOSS and enter the power loss that corresponds to the attenuation of the directional

coupler (or power splitter) at the frequency that you have entered in the previous step .

NNNNNNNNNNNNNN

Complete the power loss entry by pressing

DONE .

Remember to subtract the through arm loss from the coupler arm loss before

entering it into the power loss table , to ensure the correct power at the output

of the coupler .

4. Repeat the previous two steps to enter up to 55 frequency segments , depending on the

required accuracy .

Y ou may enter multiple segments in any order because the analyzer automatically sorts them

and lists them on the display in increasing order of frequency .

If you only enter one frequency segment, the analyzer assumes that the single value is valid

over the entire frequency range of the correction.

N

NNNNNNNNNNNNN

After you have entered all the segments , press

DONE .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PWRMTR CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PWR LOSS ON to activate the power loss compensation.

CAL

Optimizing Measurement Results 5-35

Show 17 more labels / short text on this page
Deleting Frequency Segments · · + · · press · Press · Press · Press · · · · G/n · k/m · Press · · Press ·
Page 243 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 243 · click to enlarge
Page 243

Using Sample-and-Sweep Correction Mode

Y ou can use the sample-and-sweep mode to correct the analyzer output power and update

the power meter correction data table , during the initial measurement sweep . Because the

analyzer measures the actual power at each frequency point during the initial sweep , the initial

sweep time is signi cant. However , in this mode of operation the analyzer does not require the

power meter for subsequent sweeps . Therefore , this mode sweeps considerably faster than the

continuous correction mode .

Agilent Technologies 8753D — Chapter 5 Optimizing Measurement Results diagram (page 243)
Fig. 109 · Chapter 5 Optimizing Measurement Results

Figure 5-8. Sample-and-Sweep Mode for P ower Meter Calibration

1. Calibrate and zero the power meter .
2. Connect the equipment as shown in Figure 5-8.
3. Select the HP 8753D as the system controller:

LOCAL

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SYSTEM CONTROLLER

4. Set the power meter's address:

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SET ADDRESSES

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ADDRESS: P MTR/HPIB

Select the appropriate power meter by pressing

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

POWER MTR [ ] until the correct model

number is displayed (HP 436A or HP 438A/437).

6. Set test port power to the approximate desired corrected power .

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL

PWRMTR CAL and enter the test port power level that you want at the input to

your test device . F or example , if you enter

, the display will read CAL POWER 10 .

8. If you want the analyzer to make more than one power measurement at each frequency

data point, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NUMBER OF READINGS

, (where n = the number of desired iterations).

If you increase the number of readings , the power meter correction time will substantially

increase .

5-36 Optimizing Measurement Results

Show 11 more labels / short text on this page
· · · · ## · Press · · 10 · · · n
Page 244 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 244 · click to enlarge
Page 244
9. Press

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PWRMTR CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ONE SWEEP

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TAKE CAL SWEEP .

CAL

Because power meter calibration requires a longer sweep time , you may want

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

to reduce the number of points before pressing

TAKE CAL SWEEP . After the

power meter calibration is nished, return the number of points to its original

value and the analyzer will automatically interpolate this calibration. Some

accuracy will be lost for the interpolated points .

The analyzer will use the data table for subsequent sweeps to correct the output power

level at each measurement point. Also , the status annunciator PC will appear on the

analyzer display .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Y ou can abort the calibration sweep by pressing

PWRMTR CAL OFF .

10. Remove the power sensor from the analyzer test port and connect your test device .

Using Continuous Correction Mode

Y ou can set the analyzer to update the correction table at each sweep (as in a leveling

application), using the continuous sample mode . When the analyzer is in this mode , it

continuously checks power at every point in each sweep . Y ou must keep the power meter

connected as shown in Figure 5-9. This mode is also known as power meter leveling, and the

speed is limited by the power meter .

Y ou may level at the input of a device under test, using a 2-resistor power

splitter or a directional coupler before the device; or level at the output of

the device using a 3-resistor power splitter or a bidirectional coupler after the

device .

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 244)
Fig. 110

Figure 5-9. Continuous Correction Mode for P ower Meter Calibration

1. Connect a power splitter or directional coupler to the port supplying RF power to your test

device , as shown in Figure 5-9.

2. Set test port power to approximate desired leveled power .

Optimizing Measurement Results 5-37

Page 245 of the Agilent Technologies 8753D user manual — TAKE RCVR CAL SWEEP
Page 245 · click to enlarge
Page 245

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PWRMTR CAL and enter the test port power level that you want the analyzer

CAL

to maintain at the input to your test device . Compensate for the power loss of the power

splitter or directional coupler in the setup .

4. If you want the analyzer to make more than one power measurement at each frequency

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NUMBER OF READINGS

, (where n = the number of desired

iterations).

If you increase the number of readings , the power meter correction time will substantially

increase .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

5. Press

CAL

PWRMTR CAL

EACH SWEEP

TAKE CAL SWEEP to activate the power meter

correction.

T o Calibrate the Analyzer Receiver to Measure Absolute P ower

Y ou can use the power meter calibration as a reference to calibrate the analyzer receiver to

accurately measure absolute power . The following procedure shows you how to calibrate the

receiver to any power level.

1. Set the analyzer test port power to the desired level:

N

NNNNNNNNNNNNNNNN

MENU

POWER (enter power level)

2. Connect the power sensor to the analyzer test port 1.
3. T o apply the one sweep mode , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL

PWRMTR CAL (enter power level)

ONE SWEEP

TAKE CAL SWEEP

Because power meter calibration requires a longer sweep time , you may want

to reduce the number of points before pressing

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TAKE CAL SWEEP . After the

power meter calibration is nished, return the number of points to its original

value and the analyzer will automatically interpolate this calibration.

The status notation PC will appear on the analyzer display . P ort 1 is now a calibrated source

of power .

4. Connect the test port 1 output to the test port 2 input.
5. Choose a non-ratioed measurement by pressing:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MEAS

INPUT PORTS

B

TEST PORT 1

This sets the source at PORT 1, and the measurement receiver to PORT 2, or input port B .

6. T o perform a receiver error-correction, press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL

RECEIVER CAL (enter power level)

TAKE RCVR CAL SWEEP

The receiver channel now measures power to a characteristic accuracy of 0.35 dB or better .

The accuracy depends on the match of the power meter , the source , and the receiver .

5-38 Optimizing Measurement Results

Show 17 more labels / short text on this page
Press · · data point, press · n · · · · · · · · · · · · ·
Page 246 of the Agilent Technologies 8753D user manual — Chapter 5 Optimizing Measurement Results
Page 246 · click to enlarge
Page 246

With this method, you use two precision matched adapters which are \equal." T o be equal,

the adapters must have the same match, Z

, insertion loss , and electrical delay . The adapters

in most HP calibration kits have matched electrical length, even if the physical lengths appear

di erent.

Agilent Technologies 8753D — Chapter 5 Optimizing Measurement Results diagram (page 246)
Fig. 111 · Chapter 5 Optimizing Measurement Results

Figure 5-10. Calibrating for Noninsertable Devices

T o use this method, refer to Figure 5-10 and perform the following steps:

1. P erform a transmission calibration using the rst adapter .
2. Remove adapter A, and place adapter B on port 2. A dapter B becomes the e ective test

port.

3. P erform a re ection calibration.
4. Measure the test device with adapter B in place .

The errors remaining after calibration with this method are equal to the di erences between

the two adapters that are used.

Optimizing Measurement Results 5-39

Matched A dapters

Page 247 of the Agilent Technologies 8753D user manual — F ORMA T
Page 247 · click to enlarge
Page 247

Modify the Cal Kit Thru De nition

With this method it is only necessary to use adapter B . The calibration kit thru de nition is

modi ed to compensate for the adapter and then saved as a user kit. However , the electrical

delay of the adapter must rst be found.

1. P erform a 1-port calibration on PORT 2.
2. Connect adapter B to the test port.
3. A dd a short to the open end of the B adapter .

NNNNNNNNNNNNNNNNN

Measure the delay of the adapter by pressing

F ORMA T

DELAY .

5. Divide the resulting delay measurement by 2.
6. Determine the o set delay of the calibration short by examining the de ne standard menu

(see \De ne Standard Menus").

7. Subtract the short o set delay from the value calculated in step 5. This corresponds to the

delay of adapter B .

8. Modify the calibration kit thru de nition by entering in the electrical delay of adapter B .

Save this as a user kit.

9. P erform the desired calibration with this new user kit.
10. Measure the test device .

5-40 Optimizing Measurement Results

Page 248 of the Agilent Technologies 8753D user manual — Chapter 5 Optimizing Measurement Results
Page 248 · click to enlarge
Page 248

Calibrating for Noninsertable Devices

A test device having the same sex connector on both the input and output cannot be connected

directly into a transmission test con guration. Therefore , the device is considered to be

noninsertable, and one of the following calibration methods must be performed:

modify the cal kit thru de nition

Agilent Technologies 8753D — Chapter 5 Optimizing Measurement Results diagram (page 248)
Fig. 112 · Chapter 5 Optimizing Measurement Results

Figure 5-11. Noninsertable Device

Optimizing Measurement Results 5-41

adapter removal · matched adapters

Page 249 of the Agilent Technologies 8753D user manual — Chapter 5 Optimizing Measurement Results
Page 249 · click to enlarge
Page 249

The adapter removal technique provides a means to accurately measure noninsertable devices .

The following adapters are needed:

A dapter A1, which mates with port 1 of the device , must be installed on test set port 1.

A dapter A2, which mates with port 2 of the device , must be installed on test set port 2.

A dapter A3 must match the connectors on the test device . The e ects of this adapter will be

completely removed with this calibration technique .

Agilent Technologies 8753D — Chapter 5 Optimizing Measurement Results diagram (page 249)
Fig. 113 · Chapter 5 Optimizing Measurement Results

Figure 5-12. A dapters Needed

The following requirements must also be met:

Calibration standards for performing a 2-port error correction for each connector type .

Speci ed electrical length of adapter A3 within 1/4 wavelength for the measurement

frequency range .

F or each port, a separate 2-port error correction needs to be performed to create two

calibration sets . The adapter removal algorithm uses the resultant data from the two

calibration sets and the nominal electrical length of the adapter to compute the adapter's

actual S-parameters . This data is then used to generate a separate third calibration set in

which the forward and reverse match and tracking terms are as if port 1 and port 2 could be

connected. This is possible because the actual S-parameters of the adapter are measured with

great accuracy , thus allowing the e ects of the adapter to be completely removed when the

third cal set is generated.

5-42 Optimizing Measurement Results

A dapter Removal

Page 250 of the Agilent Technologies 8753D user manual — Chapter 5 Optimizing Measurement Results
Page 250 · click to enlarge
Page 250

P erform the 2-port Error Corrections

1. Connect adapter A3 to adapter A2 on port 2. (See Figure 5-12)
Agilent Technologies 8753D — Chapter 5 Optimizing Measurement Results diagram (page 250)
Fig. 114 · Chapter 5 Optimizing Measurement Results

Figure 5-13. Two-P ort Cal Set 1

2. P erform the 2-port error correction using calibration standards appropriate for the

connector type at port 1.

When using adapter removal calibration, you must save calibration sets to the

internal disk, not to internal memory .

3. Save the results to disk. Name the le \PORT1."
4. Connect adapter A3 to adapter A1 on port 1. (See Figure 5-13)

Optimizing Measurement Results 5-43

Page 251 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 251 · click to enlarge
Page 251
Agilent Technologies 8753D — Chapter 5 Optimizing Measurement Results diagram (page 251)
Fig. 115 · Chapter 5 Optimizing Measurement Results

Figure 5-14. Two-P ort Cal Set 2

5. P erform the 2-port error correction using calibration standards appropriate for the

connector type at port 2.

6. Save the results to disk. Name the le \PORT2."
7. Determine the electrical delay of adapter A3 by performing steps 1 through 7 of \Modify

the Cal Kit Thru De nition."

Remove the A dapter

When the two sets of error correction les have been created (now referred to as \cal sets"),

the adapter may be removed.

NNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL

MORE

ADAPTER REMOVAL . This brings up the following menu:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

HELP ADAPT REMOVAL (This provides a quick reference guide to using the adapter

removal technique .)

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECALL CAL SETS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ADAPTER DELAY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ADAPTER COAX

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ADAPTER WAVEGUIDE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REMOVE ADAPTER

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECALL CAL SETS to bring up the following two choices:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECALL CAL PORT 1

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECALL CAL PORT 2

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECALL CAL SETS also brings up the internal (or external if internal not used) disk le

directory .

5-44 Optimizing Measurement Results

Press · · Press

Page 252 of the Agilent Technologies 8753D user manual — SA VE/RECALL
Page 252 · click to enlarge
Page 252

In the following two steps , calibration data is recalled, not instrument states .

10. From the disk directory , choose the le associated with the port 1 error correction, then

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECALL CAL PORT 1 .

11. When this is complete , choose the le for the port 2 error correction and press

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECALL CAL PORT 2 .

When complete , press

NNNNNNNNNNNNNNNNNNNN

RETURN .

13. Enter the value of the electrical delay of adapter A3.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ADAPTER DELAY and enter the value .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Select the appropriate key:

ADAPTER WAVEGUIDE .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REMOVE ADAPTER to complete the technique for calculating the new error

coecients and overwrite the current active calibration set in use .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

16. T o save the results of the new cal set, press

SA VE/RECALL

SELECT DISK

INTERNAL DISK

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EXTERNAL DISK )

RETURN

SAVE STATE .

Agilent Technologies 8753D — SA VE/RECALL diagram (page 252)
Fig. 116

Figure 5-15. Calibrated Measurement

V erify the Results

Since the e ect of the adapter has been removed, it is easy to verify the accuracy of the

technique by simply measuring the adapter itself. Because the adapter was used during the

creation of the two cal sets , and the technique removes its e ects , measurement of the adapter

itself should show the S-parameters .

Optimizing Measurement Results 5-45

Show 6 more labels / short text on this page
press · Press · ADAPTER COAX or · Press · · (or
Page 253 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 253 · click to enlarge
Page 253

If unexpected phase variations are observed, this indicates that the electrical delay of the

adapter was not speci ed within a quarter wavelength over the frequency range of interest.

T o correct this , recall both cal sets , since the data was previously stored to disk, change the

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

adapter delay , and press

REMOVE ADAPTER .

5-46 Optimizing Measurement Results

Page 254 of the Agilent Technologies 8753D user manual — 360 END
Page 254 · click to enlarge
Page 254

The following is an example program for performing these same operations over HP-IB:

10 ! Fileadaptrm.bas

30 ! This demonstrates how to do adapter removal over HP--IB.

50 ASSIGN @Na TO 716

70 ! Select internal disk.

110 ! Assign file #1 to the filename that has a 2-port

120 ! cal previously performed for Port 1's connector.

160 ! Recall the cal set for Port 1.

200 ! Assign file #2 to the filename that has a 2-port

210 ! cal previously performed for Port 2's connector.

250 ! Recall the cal set for Port 2.

290 ! Set the adapter electrical delay.

330 ! Perform the "remove adapter" computation.

360 END

Optimizing Measurement Results 5-47

Show 24 more labels / short text on this page
Example Program · 20 ! · 40 ! · 60 ! · 80 ! · 90 OUTPUT @Na;"INTD;" · 100 ! · 130 ! · 140 OUTPUT @Na;"TITF1""F10DCAL1"";" · 150 ! · 170 ! · 180 OUTPUT @Na;"CALSPORT1;" · 190 ! · 220 ! · 230 OUTPUT @Na;"INTD;TITF2""F20DCAL2"";" · 240 ! · 260 ! · 270 OUTPUT @Na;"CALSPORT2;" · 280 ! · 300 ! · 310 OUTPUT @Na;"ADAP158PS;" · 320 ! · 340 ! · 350 !OUTPUT @Na;"MODS;"
Page 255 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 255 · click to enlarge
Page 255

Making A ccurate Measurements of Electrically Long Devices

A device with a long electrical delay , such as a long length of cable or a SA W lter , presents

some unusual measurement problems to a network analyzer operating in swept frequency

mode . Often the measured response is dependent on the analyzer's sweep time , and incorrect

data may be obtained. At faster sweep rates , the magnitude of the response may seem to drop

and look distorted, while at slower sweep rates it looks correct. The results may indicate that a

cable has more loss than it truly does , or that a lter has some unusual ripple in the passband

which isn't really there .

This section describes the cause of this behavior , and how to accurately measure these

electrically long devices .

The Cause of Measurement Problems

When using a vector network analyzer to measure a device that has a long electrical delay

(T), the device's time delay causes a frequency shift between its input and output signals . The

frequency shift, F , equals the product of the sweep rate and the time delay:

F= dF/dt * T

Since frequency is changing with time as the analyzer sweeps , the time delay of the DUT

causes a frequency o set between its input and output. In the analyzer receiver , the test and

reference input signals will di er in frequency by F . Because the test signal frequency is

slightly di erent than the receiver frequency , the analyzer will err in measuring its magnitude

or phase . The faster the analyzer's sweep rate , the larger F becomes , and the larger the error

in the test channel.

The HP 8753D network analyzers do not sweep at a constant rate . The frequency range is

covered in several bands , and the sweep rate may be di erent in each band. So if an operator

sets up a broadband sweep with the minimum sweep time , the error in measuring a long device

will be di erent in each band, and the data will be discontinuous at each band edge . This can

produce confusing results which make it dicult to determine the true response of the device .

T o Improve Measurement Results

T o reduce the error in these measurements , the frequency shift, F , must be reduced. F can

be reduced by using the following three methods:

decreasing the sweep rate

decreasing the time delay (T)

Decreasing the Sweep Rate

The sweep rate can be decreased by increasing the analyzer's sweep time . T o increase the

analyzer's sweep time , press

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MENU

SWEEP TIME [MANUAL] and use the front panel knob , the

keys , or the front panel keypad enter in the appropriate sweep time .

Selection of the appropriate sweep time depends on the device being measured; the longer the

electrical delay of the device under test, the slower the sweep rate must be . A good way to tell

when the sweep rate is slow enough is to put the vector network analyzer into a list frequency

mode of sweeping, and compare the data. In this mode , the vector network analyzer does not

sweep the frequency , but steps to each listed frequency point, stops , makes a measurement,

then goes on to the next point. Because errors do not occur in the list frequency mode , it can

be used to check the data. The disadvantage of the list frequency mode is that it is slower than

sweeping.

5-48 Optimizing Measurement Results

Show 4 more labels / short text on this page
· step · * · +
Page 256 of the Agilent Technologies 8753D user manual — Chapter 5 Optimizing Measurement Results
Page 256 · click to enlarge
Page 256

Decreasing the Time Delay

The other way to reduce F is by decreasing the time delay , T . Since T is a property of

the device that is being measured, it cannot literally be decreased. However , what can be

decreased is the di erence in delay times between the paths to the R channel and the B

channel. These times can be equalized by adding a length of cable to the R channel which has

approximately the same delay as the device under test.

This length of cable can be inserted between the R CHANNEL IN and OUT connectors on the

front panel of the analyzer . The delay of this cable must be less than 5s .

Optimizing Measurement Results 5-49

Page 257 of the Agilent Technologies 8753D user manual — Chapter 5 Optimizing Measurement Results
Page 257 · click to enlarge
Page 257

Y ou can increase the analyzer sweep speed by avoiding the use of some features that require

computational time for implementation and updating, such as bandwidth marker tracking.

Y ou can also increase the sweep speed by making adjustments to the measurement settings .

The following suggestions for increasing sweep speed are general rules that you should

experiment with:

decrease the frequency span

set the auto sweep time mode

widen the system bandwidth

reduce the averaging factor

reduce the number of measurement points

set the sweep type

use chop sweep mode

fast 2-port calibration mode

T o Decrease the Frequency Span

The hardware of the network analyzer sweeps the frequency range in separate bands , where

switching from band to band takes time . Modify the frequency span to eliminate as many band

switches as possible while maintaining measurement integrity . Refer to the following table to

identify the analyzer's band switch points:

T able 5-5. Band Switch P oints

0 .01 MHz to .3 MHz

1 .3 MHz to 3.3 MHz

2 3.3 MHz to 16 MHz

3 16 MHz to 31 MHz

4 31 MHz to 61 MHz

5 61 MHz to 121 MHz

6 121 MHz to 178 MHz

7 178 MHz to 296 MHz

8 296 MHz to 536 MHz

9 536 MHz to 893 MHz

10 893 MHz to 1.607 GHz

11 1.607 GHz to 3 GHz

12 (Option 006) 3 GHz to 4.95 GHz

13 (Option 006) 4.95 GHz to 6 GHz

5-50 Optimizing Measurement Results

Increasing Sweep Speed · use external calibration · Band Frequency Span

Page 258 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 258 · click to enlarge
Page 258

T o Set the A uto Sweep Time Mode

A uto sweep time mode is the default mode (the preset mode). This mode maintains the fastest

sweep speed possible for the current measurement settings .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MENU

SWEEP TIME

, to re-enter the auto mode .

Optimizing Measurement Results 5-51

Press · · 0

Page 259 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 259 · click to enlarge
Page 259

T o Widen the System Bandwidth

NNNNNNNNNNNNNNNNN

IF BW .

A V G

2. Set the IF bandwidth to change the sweep time .

The following table shows the relative increase in sweep time as you decrease system

bandwidth. The characteristic values in the following table were derived using 201

measurement points .

IF B W Cycle Time (Seconds)

Full Span Narrow Sweep

3700 Hz .446 .150

3000 Hz .447 .176

1000 Hz .511 .312

300 Hz .944 .980

100 Hz 2.25 2.070

30 Hz 7.57 7.240

10 Hz 21.98 21.600

1 The listed sweep times correspond

to an HP 8753D analyzer being set

to a preset state for the full span,

and 2 GHz to 3 GHz for the narrow

span.

T o Reduce the A veraging F actor

By reducing the averaging factor (number of sweeps) or switching o averaging, you can

increase the analyzer's measurement speed. The time needed to compute averages can also

slow the sweep time slightly , in narrow spans .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

A V G

AVG FACTOR .

2. Enter an averaging factor that is less than the value displayed on the analyzer screen and

.

If you want to switch o averaging, press

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AVERAGING OFF .

A V G

T o Reduce the Number of Measurement P oints

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MENU

NUMBER OF POINTS .

2. Enter a number of points that is less than the value displayed on the analyzer screen and

.

5-52 Optimizing Measurement Results

Show 9 more labels / short text on this page
Press · · Press · · press · · Press · · press
Page 260 of the Agilent Technologies 8753D user manual — N
Page 260 · click to enlarge
Page 260

The analyzer sweep time does not change proportionally with the number of points , but as

indicated below .

Full Span Narrow Span

LIN LIST/LOG LIN LIST

51 0.35 0.57 0.09 0.25

101 0.39 0.77 0.12 0.43

201 0.43 1.11 0.17 0.78

401 0.49 1.73 0.27 1.33

801 0.69 3.04 0.47 2.64

1601 1.09 5.7 0.87 5.30

1 The listed sweep times correspond to the

analyzer being set to a preset state , with a

6 GHz span. A 3 GHz span would have

faster sweep times .

T o Set the Sweep Type

Di erent sweep speeds are associated with the following three types of non-power sweeps .

Choose the sweep type that is most appropriate for your application.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MENU

SWEEP TYPE MENU .

2. Select the sweep type:

NNNNNNNNNNNNNNNNNNNNNNNNNN

LIN FREQ for the fastest sweep for a given number of xed points .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LIST FREQ for the fastest sweep when speci c non-linearly spaced frequency

points are of interest.

NNNNNNNNNNNNNNNNNNNNNNNNNN

LOG FREQ for the fastest sweep when the frequency points of interest are in the

lower part of the frequency span selected.

T o View a Single Measurement Channel

Viewing a single channel will increase the measurement speed if the analyzer's channels are in

alternate , or uncoupled mode .

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DISPLA Y

DUAL CHAN OFF .

2. Press

CHAN 1 and

CHAN 2 to alternately view the two measurement channels .

If you must view both measurement channels simultaneously (with dual channel), use the

chop sweep mode , explained next.

Optimizing Measurement Results 5-53

Show 10 more labels / short text on this page
Number of · Sweep Time (Seconds) · P oints · Press · · Select · Select · Select · Press · ]
Page 261 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 261 · click to enlarge
Page 261

T o A ctivate Chop Sweep Mode

Y ou can use the chop sweep mode to make two measurements at the same time . F or example ,

the analyzer can measure A/R and B/R simultaneously . Y ou can activate the chop mode by

PRESET

or by the following the sequence below .

CAL

NNNNNNNNNNNNNN

MORE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CHOP A and B .

F or more information, refer to \Alternate and Chop Sweep Modes" in Chapter 6.

T o Use External Calibration

Ooading the error correction process to an external PC increases throughput on the network

analyzer . This can be accomplished with remote only commands . Refer to the HP 8753D

Programmer's Guide for information on how to use external calibration.

T o Use F ast 2-P ort Calibration

With the 2-port calibration on, faster measurements may be made by not measuring the reverse

path for every forward sweep . This is controlled by the test set switch command.

When making measurements using full two-port error-correction, the following types of test set

switching can be de ned by the user:

Hold: In this mode the analyzer does not switch between the test ports on every sweep . The

measurement stays on the active port after an initial cycling between the ports . The fastest

measurements can be made by using this type of test set switching. Pressing the

MEAS

or changing to a di erent S-parameter measurement will cause the test set to switch and

cycle between the ports .

Continuous: In this mode the analyzer will switch between the test ports on every sweep .

Although this type of test set switching provides the greatest measurement accuracy , it also

takes the longest amount of time .

Number of SweepsIn this mode there is an initial cycling between the test ports and

then the measurement stays on the active port for a user-de ned number of sweeps . After

the speci ed number of sweeps have been executed, the analyzer switches between the

test ports and begins the cycle again. This type of test set switching can provide improved

measurement accuracy over the hold mode and faster measurement speeds than continuous

mode .

1. T o access the test set switch functions , press:

NNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL

MORE

TEST SET SW

2. T o activate the hold mode , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

The analyzer will then display

TEST SET SW HOLD .

3. T o activate the continuous mode , press:

The analyzer will then display

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TEST SET SW CONTINUOUS .

5-54 Optimizing Measurement Results

Show 12 more labels / short text on this page
pressing · Press · · key · · · · 0 · · · 1 ·
Page 262 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 262 · click to enlarge
Page 262
4. T o enter the number of sweeps , press:

The analyzer will then display

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TEST SET SW X SWEEPS .

Optimizing Measurement Results 5-55

· ·

Page 263 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 263 · click to enlarge
Page 263

Dynamic range is the di erence between the analyzer's maximum allowable input level and

minimum measurable power . F or a measurement to be valid, input signals must be within these

boundaries . The dynamic range is a ected by these factors:

test port input power

test port noise oor

T o Increase the T est P ort Input P ower

Y ou can increase the analyzer's source output power so that the test device output power is at

the top of the measurement range of the analyzer test port.

NNNNNNNNNNNNNNNNN

POWER and enter the new source power level, followed by

.

MENU

TEST PORT INPUT D AMA GE LEVEL: +26 dBm

T o Reduce the Receiver Noise Floor

Since the dynamic range is the di erence between the analyzer's input level and its noise oor ,

using the following techniques to lower the noise oor will increase the analyzer's dynamic

range .

Each tenfold reduction in IF (receiver) bandwidth lowers the noise oor by 10 dB . F or example ,

changing the IF bandwidth from 3 kHz to 300 Hz, you will lower the noise oor by about

10 dB .

1. Press

NNNNNNNNNNNNNNNNN

IF BW .

A V G

2. Enter the bandwidth value that you want, followed by

.

Changing Measurement A veraging

Y ou can apply weighted averaging of successive measurement traces to remove the e ects of

random noise .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

1. Press

A V G

AVERAGING FACTOR .

2. Enter a value followed by

.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AVERAGING ON .

Refer to the \Application and Operation Concepts" chapter for more information on

averaging.

5-56 Optimizing Measurement Results

Show 8 more labels / short text on this page
Increasing Dynamic Range · receiver crosstalk · Press · · Changing System Bandwidth · · · Press
Page 264 of the Agilent Technologies 8753D user manual — CAL
Page 264 · click to enlarge
Page 264

Y ou can use two analyzer functions to help reduce the e ect of noise on the data trace:

T o A ctivate A veraging

The noise is reduced with each new sweep as the e ective averaging factor increments .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

A VG

AVERAGING FACTOR .

2. Enter a value followed by

.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AVERAGING ON .

Refer to the \Application and Operation Concepts" chapter for more information on

averaging.

T o Change System Bandwidth

By reducing the system bandwidth, you reduce the noise that is measured during the sweep .

While averaging requires multiple sweeps to reduce noise , narrowing the system bandwidth

reduces the noise on each sweep .

NNNNNNNNNNNNNNNNN

A VG

IF BW .

2. Enter the IF bandwidth value that you want, followed by

.

Narrower system bandwidths cause longer sweep times . When in auto sweep time mode , the

analyzer uses the fastest sweep time possible for any selected system bandwidth. A uto sweep

time mode is the default (preset) analyzer setting.

T o reduce receiver crosstalk you can do the following:

perform a response and isolation measurement calibration

set the sweep to the alternate mode

Alternate sweep is intended for measuring wide dynamic range devices , such as high pass and

bandpass lters . This sweep mode removes a type of leakage term through the device under

test, from one channel to another .

T o set the alternate sweep , press

NNNNNNNNNNNNNN

MORE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ALTERNATE A AND B .

CAL

Refer to the procedures , located earlier in this chapter for a response and isolation

measurement calibration procedure .

Optimizing Measurement Results 5-57

Show 10 more labels / short text on this page
Reducing Trace Noise · activate measurement averaging · reduce system bandwidth · Press · · Press · Press · · Reducing Receiver Crosstalk ·
Page 265 of the Agilent Technologies 8753D user manual — N
Page 265 · click to enlarge
Page 265

T o reduce time during recall and frequency changes , the raw o set function and the spur

avoidance function can be turned o . T o turn these functions o , press:

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SYSTEM

CONFIGURE MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RAW OFFSET OFF

SPUR AVOID OFF

The raw o set function is normally on and controls the sampler and attenuator o sets . The

spur avoidance function is normally on and generates values as part of the sampler o set table .

The creation of this table takes considerable time during a recall of an instrument state .

T o save time at recalls and during frequency changes , both functions should be turned o . This

will avoid generating the sampler o set table .

Raw o sets may be turned on or o individually for each channel. They follow the channel

coupling. F or dual channel operation, raw o sets should be turned o for each channel if the

channels are uncoupled. Spur avoidance is always coupled between channels , therefore both

channels are turned on or o at the same time .

Both functions must be turned o to realize the recall time savings .

The following table lists the recall state times with the raw o sets and spur avoidance

functions on or o .

T able 5-6. Typical Recall State Times

Operations Channel P oints Raw O set

Recall and Sweep Dual Chan. 1601 On 3.89 3.18

Recall and Sweep Dual Chan. 1601 O 2.008 1.298

Dual Chan. 1601 n/a 0.71 no recall

Recall and Sweep Dual Chan. 201 On 0.955 .740

Recall and Sweep Dual Chan. 201 O 0.734 .519

Dual Chan. 201 n/a 0.215 no recall

Recall and Sweep Single Chan. 1601 On 2.134 1.424

Recall and Sweep Single Chan. 1601 O 1.251 .541

Instrument StateCF= 1GHz, Span= 2MHz, Error Correction OFF . HP-IB commands sent for timing are

Recall;OPC?;SING; or , for sweep only , OPC?;SING;.

5-58 Optimizing Measurement Results

Show 12 more labels / short text on this page
Reducing Recall Time · · · T otal Time · Recall Time · Spur A void · (secs) · (secs) · Sweep only (no · Recall) · Sweep only (no · Recall)
Page 266 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 266 · click to enlarge
Page 266

Understanding Spur A voidance

In the 400 MHz to 3 GHz range , where the source signal is created by heterodyning two higher

frequency oscillators , unwanted spurious mixing products from the source may be present at

the output. These spurs can become apparent in lter measurements when lters have greater

than 80 dB rejection.

Spur avoidance slightly moves the frequency of both oscillators such that the source frequency

remains the same but the spurious mixing products shift out of the measurement receiver

range . The calculation of the exact frequency points where the shifting must occur (stored

in the sampler o set table) increases the time needed to change or recall instrument states .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SPUR AVOID OFF and

RAW OFFSET OFF eliminates this calculation.

Optimizing Measurement Results 5-59

Selecting

Page 267 of the Agilent Technologies 8753D user manual — Chapter 5 Optimizing Measurement Results
Page 267 · click to enlarge
Page 267
Page 267 of the Agilent Technologies 8753D user manual — Chapter 5 Optimizing Measurement Results
Page 268 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 268 · click to enlarge
Page 268

Application and Operation Concepts

This chapter provides conceptual information on the following primary operations and

applications that are achievable with the HP 8753D network analyzer .

HP 8753D System operation

A ctive channel keys

Where to Look for More Information

A dditional information about many of the topics discussed in this chapter is located in the

following areas:

Chapter 2, \Making Measurements ," contains step-by-step procedures for making

measurements or using particular functions .

Chapter 3, \Mixer Measurements ," contains step-by-step procedures for making

measurements of mixers .

Chapter 5, \Optimizing Measurement Results ," describes techniques and functions for

achieving the best measurement results .

Chapter 7, \Speci cations and Measurement Uncertainties ," de nes the performance

capabilities of the analyzer .

Chapter 8, \Menu Maps ," shows softkey menu relationships .

Chapter 9, \Key De nitions ," describes all the front panel keys and softkeys .

Application and Operation Concepts 6-1

Show 18 more labels / short text on this page
Data processing · Entry block keys · Stimulus functions · Response functions · S-parameters · Display formats · Scale reference · Display functions · A veraging · Markers · Measurement calibration · Instrument state functions · Time domain operation · T est sequencing · Ampli er measurements · Mixer measurements · Connection considerations · Reference documents
Page 269 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 269 · click to enlarge
Page 269

HP 8753D System Operation

Network analyzers measure the re ection and transmission characteristics of devices and

networks . A network analyzer test system consists of the following:

The analyzer applies a signal that is transmitted through the test device , or re ected from its

input, and then compares it with the incident signal generated by the swept RF source . The

signals are then applied to a receiver for measurement, signal processing, and display .

The HP 8753D vector network analyzer integrates a high resolution synthesized RF source ,

test set, and a dual channel three-input receiver to measure and display magnitude , phase ,

and group delay of transmitted and re ected power . With Option 010, the analyzer has the

additional capability of transforming measured data from the frequency domain to the time

domain. Other options are explained in Chapter 1, \HP 8753D Description and Options ."

Figure 6-1 is a simpli ed block diagram of the network analyzer system. A detailed block

diagram of the analyzer is provided in the HP 8753D Network Analyzer Service Guide together

with a theory of system operation.

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 269)
Fig. 117 · Chapter 6 Application and Operation Concepts

Figure 6-1. Simpli ed Block Diagram of the Network Analyzer System

The Built-In Synthesized Source

The analyzer's built-in synthesized source produces a swept RF signal or CW (continuous wave)

signal in the range of 30 kHz to 3.0 GHz. The HP 8753D Option 006 is able to generate signals

up to 6 GHz. The RF output power is leveled by an internal ALC (automatic leveling control)

circuit. T o achieve frequency accuracy and phase measuring capability , the analyzer is phase

locked to a highly stable crystal oscillator . F or this purpose , a portion of the transmitted signal

is routed to the R channel input of the receiver , where it is sampled by the phase detection

loop and fed back to the source . Some portion of the RF source signal must always be sent to

the R channel input. The level must be between 0 dB and -35 dBm.

The Source Step Attenuator

The step attenuator contained in the source is used to adjust the power level to the test device

without changing the level of the incident power in the reference path.

6-2 Application and Operation Concepts

Show 4 more labels / short text on this page
source · signal-separation devices · receiver · display
Page 270 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 270 · click to enlarge
Page 270

The Built-In T est Set

The HP 8753D features a built-in test set that provides connections to the test device , as well

as to the signal-separation devices . The signal separation devices are needed to separate the

incident signal from the transmitted and re ected signals . The incident signal is applied to the

R channel input through a jumper cable on the front panel. Meanwhile , the transmitted and

re ected signals are internally routed from the test port couplers to the inputs of the A and B

sampler/mixers in the receiver . P ort 1 is connected to the A input and port 2 is connected to

the B input.

The test set contains the hardware required to make simultaneous transmission and re ection

measurements in both the forward and reverse directions . An RF path switch in the test set

allows reverse measurements to be made without changing the connections to the test device .

The receiver block contains three sampler/mixers for the R, A, and B inputs . The signals

are sampled, and mixed to produce a 4 kHz IF (intermediate frequency). A multiplexer

sequentially directs each of the three signals to the ADC (analog to digital converter)

where it is converted from an analog to a digital signal. The signals are then measured and

processed for viewing on the display . Both amplitude and phase information are measured

simultaneously , regardless of what is displayed on the analyzer .

A microprocessor takes the raw data and performs all the required error correction, trace

math, formatting, scaling, averaging, and marker operations , according to the instructions

from the front panel or over HP-IB . The formatted data is then displayed. The data processing

sequence is described in \Data Processing" later in this chapter .

Required P eripheral Equipment

Measurements require calibration standards for vector accuracy enhancement (error-

correction), and cables for interconnections . Model numbers and details of compatible power

splitters , calibration kits , and cables are provided in Chapter 11, \Compatible P eripherals ."

Application and Operation Concepts 6-3

The Receiver Block · The Microprocessor

Page 271 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 271 · click to enlarge
Page 271

The analyzer's receiver converts the R, A, and B input signals into useful measurement

information. This conversion occurs in two main steps:

The swept high frequency input signals are translated to xed low frequency IF signals , using

analog sampling or mixing techniques . (Refer to the HP 8753D Network Analyzer Service

Guide for more details on the theory of operation.)

The IF signals are converted into digital data by an analog to digital converter (ADC). From

this point on, all further signal processing is performed mathematically by the analyzer

microprocessors .

The following paragraphs describe the sequence of math operations and the resulting data

arrays as the information ows from the ADC to the display . They provide a good foundation

for understanding most of the response functions , and the order in which they are performed.

Figure 6-2 is a data processing ow diagram that represents the ow of numerical data from IF

detection to display . The data passes through several math operations , denoted in the gure by

single line boxes . Most of these operations can be selected and controlled with the front panel

response block menus . The data, stored in arrays along the way and denoted by double line

boxes , are places in the ow path where data is accessible via HP-IB .

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 271)
Fig. 118 · Chapter 6 Application and Operation Concepts

Figure 6-2. Data Processing Flow Diagram

6-4 Application and Operation Concepts

Data Processing

Page 272 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 272 · click to enlarge
Page 272

While only a single ow path is shown, two identical paths are available , corresponding to

channel 1 and channel 2. When the channels are uncoupled, each channel is processed and

controlled independently .

Data point de nitionA \data point" or \point" is a single piece of data representing a

measurement at a single stimulus value . Most data processing operations are performed

point-by-point; some involve more than one point.

Sweep de nitionA \sweep" is a series of consecutive data point measurements , taken over

a sequence of stimulus values . A few data processing operations require that a full sweep

of data is available . The number of points per sweep can be de ned by the user . The units

of the stimulus values (such as power , frequency , and time) can change , depending on the

sweep mode , although this does not generally a ect the data processing path.

The ADC (analog-to-digital converter) converts the R, A, and B inputs (already down-converted

to a xed low frequency IF) into digital words . (The A UX INPUT connector on the rear panel is

a fourth input.) The ADC switches rapidly between these inputs , so they are converted nearly

simultaneously .

This detection occurs in the digital lter , which performs the discrete F ourier transform

(DFT) on the digital words . The samples are converted into complex number pairs (real plus

imaginary , R+jX). The complex numbers represent both the magnitude and phase of the IF

signal. If the A UX INPUT is selected, the imaginary part of the pair is set to zero . The DFT

lter shape can be altered by changing the IF bandwidth, which is a highly e ective technique

for noise reduction.

These calculations are performed if the selected measurement is a ratio of two inputs (for

example , A/R or B/R). This is a complex divide operation. If the selected measurement is

absolute (such as A or B), no calculations are performed. The R, A, and B values are also split

into channel data at this point.

The next digital processing technique used is sampler/IF correction. This process digitally

corrects for frequency response errors (both magnitude and phase , primarily sampler rollo ) in

the analog down-conversion path.

Sweep-T o-Sweep A veraging

A veraging is another noise reduction technique . This calculation involves taking the complex

exponential average of several consecutive sweeps . This technique cannot be used with

single-input measurements .

Application and Operation Concepts 6-5

Show 5 more labels / short text on this page
Processing Details · The ADC · IF Detection · Ratio Calculations · Sampler/IF Correction
Page 273 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 273 · click to enlarge
Page 273

These data arrays store the results of all the preceding data processing operations . (Up to

this point, all processing is performed real-time with the sweep by the IF processor . The

remaining operations are not necessarily synchronized with the sweep , and are performed

by the main processor .) When full 2-port error correction is on, the raw arrays contain all

four S-parameter measurements required for accuracy enhancement. When the channels are

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

COUPLED CH OFF ), there may be as many as eight raw arrays . These arrays are

directly accessible via HP-IB . Notice that the numbers here are still complex pairs .

Raw arrays contain the pre-raw data which has sampler and attenuator o set applied.

V ector Error-correction (A ccuracy Enhancement)

Error-correction is performed next, if a measurement calibration has been performed and

correction is activated. Error-correction removes repeatable systematic errors (stored in the

error coecient arrays) from the raw arrays . This can vary from simple vector normalization to

full 12-term error-correction.

The results of error-correction are stored in the data arrays as complex number pairs . These

are subsequently used whenever correction is on, and are accessible via HP-IB .

If the data-to-memory operation is performed, the data arrays are copied into the memory

arrays .

This operation selects either the data array , memory array , or both to continue owing through

the data processing path. In addition, the complex ratio of the two (data/memory) or the

di erence (datamemory) can also be selected. If memory is displayed, the data from the

memory arrays goes through exactly the same processing ow path as the data from the data

arrays .

Gating (Option 010 Only)

This digital ltering operation is associated with time domain transformation. Its purpose

is to mathematically remove unwanted responses isolated in time . In the time domain, this

can be viewed as a time-selective bandpass or bandstop lter . (If both data and memory are

displayed, gating is applied to the memory trace only if gating was on when data was stored

into memory .)

The Electrical Delay Block

This block involves adding or subtracting phase in proportion to frequency . This is equivalent

to \line-stretching" or arti cially moving the measurement reference plane . This block also

includes the e ects of port extensions as well as electrical delay .

This converts the measured S-parameter data to the equivalent complex impedance (Z) or

admittance (Y) values , or to inverse S-parameters (1/S).

Transform (Option 010 Only)

This transform converts frequency domain information into the time domain when it is

activated . The results resemble time domain re ectometry (TDR) or impulse-response

measurements . The transform uses the chirp-Z inverse fast F ourier transform (FFT) algorithm

to accomplish the conversion. The windowing operation, if enabled, is performed on the

6-6 Application and Operation Concepts

Show 5 more labels / short text on this page
Pre-Raw Data Arrays · uncoupled ( · Raw Arrays · Trace Math Operation · Conversion
Page 274 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 274 · click to enlarge
Page 274

frequency domain data just before the transform. (A special transform mode is available

to \demodulate" CW sweep data, with time as the stimulus parameter , and display spectral

information with frequency as the stimulus parameter .)

This operation converts the complex number pairs into a scalar representation for display ,

according to the selected format. This includes group delay calculations . These formats are

often easier to interpret than the complex number representation. (P olar and Smith chart

formats are not a ected by the scalar formatting.) Notice in Figure 6-2 that after formatting, it

is impossible to recover the complex data.

This noise reduction technique smoothes noise on the trace . Smoothing is also used to set the

aperture for group delay measurements .

When smoothing is on, each point in a sweep is replaced by the moving average value of

several adjacent (formatted) points . The number of points included depends on the smoothing

aperture , which can be selected by the user . The e ect is similar to video ltering. If data and

memory are displayed, smoothing is performed on the memory trace only if smoothing was on

when data was stored into memory .

The data processing results are now stored in the format arrays . Notice in Figure 6-2 that the

marker values and marker functions are all derived from the format arrays . Limit testing is

also performed on the formatted data. The format arrays are accessible via HP-IB .

O set and Scale

These operations prepare the formatted data for display . This is where the reference line

position, reference line value , and scale calculations are performed, as appropriate to the

format.

The display memory stores the display image for presentation on the analyzer . The information

stored includes graticules , annotation, and softkey labels . If user display graphics are written,

these are also stored in display memory . When a print or plot is made , the information is taken

from display memory .

The display is updated frequently and synchronously with the data processing operations .

Application and Operation Concepts 6-7

Show 4 more labels / short text on this page
F ormat · Smoothing · F ormat Arrays · Display Memory
Page 275 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 275 · click to enlarge
Page 275

A ctive Channel K eys

The analyzer has two digital channels for independent measurements . Two di erent sets

of data can be measured simultaneously , for example , the re ection and transmission

characteristics of a device , or one measurement with two di erent frequency spans . The

analyzer can separately , or simultaneously , show the data.

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 275)
Fig. 119 · Chapter 6 Application and Operation Concepts

Figure 6-3. A ctive Channel K eys

CHAN 1 and

CHAN 2 keys shown in Figure 6-3 allow you to select the \active

channel." The front panel keys currently allow you to control the active channel. All of the

channel-speci c functions that you select apply to the active channel. The current active

channel is indicated by an amber LED adjacent to the corresponding channel key .

The analyzer has dual channel capability , so that you can view both the active and inactive

channel traces , either overlaid or on separate graticules one above the other (split display).

The dual channel and split display features are available in the display menus . Refer to

\Display Menu" later in this chapter for illustrations and descriptions of the di erent display

capabilities .

Uncoupling Stimulus V alues Between Channels

Y ou can uncouple the stimulus values between the two display channels by pressing

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

COUPLED CH OFF . This allows you to assign di erent stimulus values for each channel; it's

almost like having the use of a second analyzer . The coupling and uncoupling of the stimulus

values for the two channels are independent of the display and marker functions . Refer to

\Channel Stimulus Coupling" later in this chapter for a listing of the stimulus parameters

associated with the coupled channel mode .

Measurement markers can have the same stimulus values (coupled) for the two channels , or

they can be uncoupled for independent control in each channel. Refer to \Markers" later in

this chapter for more information about markers .

6-8 Application and Operation Concepts

The · Dual Channel · Coupled Markers

Page 276 of the Agilent Technologies 8753D user manual — M/
Page 276 · click to enlarge
Page 276

Entry Block K eys

The entry block, illustrated in Figure 6-4, includes the numeric and units keypad, the knob ,

and the step keys . Y ou can use these in combination with other front panel keys and softkeys

for the following reasons:

to modify the active entry

to enter or change numeric data

to change the value of the active marker

Generally , the keypad, knob , and step keys can be used interchangeably .

Before you can modify a function, you must activate the particular function by pressing the

corresponding front panel key or softkey . Then you can modify the value directly with the

knob , the step keys , or the digits keys and a terminator .

If no other functions are activated, the knob moves the active marker .

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 276)
Fig. 120 · Chapter 6 Application and Operation Concepts

Figure 6-4. Entry Block

The units terminator keys are the four keys in the right column of the keypad. Y ou must

use these keys to specify units of numerical entries from the keypad. A numerical entry is

incomplete until a terminator is supplied. The analyzer indicates that an input is incomplete

by a data entry arrow pointing at the last entered digit in the active entry area. When you

press the units terminator key , the arrow is replaced by the units you selected. The units are

abbreviated on the terminator keys as follows:

M/

= basic units: dB , dBm, degrees , seconds , Hz, or dB/GHz (may be used to terminate

unitless entries such as averaging factor)

Application and Operation Concepts 6-9

Show 22 more labels / short text on this page
Units T erminator · · = · Giga/nano (10 · / 10 · 9 · ) · G/n · · = · Mega/micro (10 · / 10 · 6 · ) · · = · 3 · k/m · kilo/milli (10 · / 10 · ) ·
Page 277 of the Agilent Technologies 8753D user manual — ST ART
Page 277 · click to enlarge
Page 277

Y ou can use the knob to make continuous adjustments to current measurement parameter

values or the active marker position. V alues changed by the knob are e ective immediately ,

and require no units terminator .

Y ou can use the step keys

* (up) and

(down) to step the current value of the active

function up or down. The analyzer de nes the steps for di erent functions . No units

terminator is required. F or editing a test sequence , you can use these keys to scroll through the

displayed sequence .

ENTRY OFF

Y ou can use this key to clear and turn o the active entry area, as well as any displayed

prompts , error messages , or warnings . Use this function to clear the display before printing or

plotting. This key also helps prevent changing active values accidentally by moving the knob .

Y ou can use this key to delete the last entry , or the last digit entered from the numeric keypad.

Y ou can also use this key in one of two ways for modifying a test sequence:

deleting a single-key command that you may have pressed by mistake (for example

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Trans:FWD S21 (B/R) )

deleting the last digit in a series of digits that you may have input, as long as you haven't yet

pressed a terminator (for example if you pressed

ST ART

12 but did not press

Y ou can use this key to add a decimal point to the number you entered.

Y ou can use this key to add a minus sign to the number you entered.

6-10 Application and Operation Concepts

Show 16 more labels / short text on this page
Knob · Step K eys · + · · · · · · G/n · , etc) · · · · · ·
Page 278 of the Agilent Technologies 8753D user manual — SP AN
Page 278 · click to enlarge
Page 278
Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 278)
Fig. 121 · Chapter 6 Application and Operation Concepts

Figure 6-5. Stimulus Function Block

The stimulus function block keys are used to de ne the source RF output signal to the test

device by providing control of the following parameters:

time domain start and stop times (Option 010 Only)

power sweep start and stop values

RF power level and power ranges

number of data points

channel and test port coupling

De ning Ranges with Stimulus K eys

ST ART

STOP

CENTER

, and

SP AN

keys are used to de ne the swept frequency range ,

time domain range (Option 010), or power sweep range . The range can be expressed as either

start/stop or center/span. When one of these keys is pressed, its function becomes the active

function. The value is displayed in the active entry area and can be changed with the knob ,

step keys , or numeric keypad. Current stimulus values for the active channel are also displayed

along the bottom of the graticule . Frequency values can be blanked for security purposes ,

using the display menus .

The preset stimulus mode is frequency , and the start and stop stimulus values are set to 30

kHz and 3 GHz (or 6 GHz with Option 006) respectively . In the time domain (Option 010) or in

CW time mode , the stimulus keys refer to time (with certain exceptions). In power sweep , the

stimulus value is in dBm.

Because the display channels are independent, the stimulus signals for the two channels can be

uncoupled and their values set independently . The values are then displayed separately if the

instrument is in dual channel display mode . In the uncoupled mode with dual channel display

the instrument takes alternate sweeps to measure the two sets of data. Channel stimulus

Application and Operation Concepts 6-11

Show 9 more labels / short text on this page
Stimulus Functions · swept frequency ranges · sweep time · sweep trigger · CW frequency · sweep type · The · , · ,
Page 279 of the Agilent Technologies 8753D user manual — N
Page 279 · click to enlarge
Page 279

coupling is explained in the \Stimulus Menu" section, and dual channel display capabilities are

explained in the \Display Menu" section located later in this chapter .

MENU

key provides access to the stimulus menu, which consists of softkeys that activate

stimulus functions or provide access to additional menus . These softkeys are used to de ne and

control all stimulus functions other than start, stop , center , and span. The following softkeys

are located within the stimulus menu:

NNNNNNNNNNNNNNNNN

POWER provides access to the power menu.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SWEEP TIME allows you to specify the sweep time .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRIGGER MENU provides access to the trigger menu.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NUMBER of POINTS allows you to specify the number of measurement points per sweep .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MEASURE RESTART allows you to cause the current measurement to abort and a new

measurement to begin. With two-port error-correction activated, pressing this softkey causes

all four S-parameters to be remeasured.

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

COUPLED CH ON off allows you to couple or uncouple the stimulus functions of the two

display channels .

NNNNNNNNNNNNNNNNNNNNNNN

CW FREQ allows you to specify the CW frequency .

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SWEEP TYPE MENU provides access to the sweep type menu.

6-12 Application and Operation Concepts

Stimulus Menu · The

Page 280 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 280 · click to enlarge
Page 280

The P ower Menu

The power menu is used to de ne and control analyzer power . It consists of the following

softkeys:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PWR RANGE AUTO man allows you to select power ranges automatically or manually .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

POWER RANGES provides access to the power range menu.

NNNNNNNNNNNNNNNNN

SLOPE compensates for power loss versus the frequency sweep , by sloping the output power

upwards proportionally to frequency .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SLOPE on OFF toggles the power slope function on or o .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SOURCE PWR ON off allows you to switch the source power on or o . When a power trip

occurs , the trip is reset by selecting

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SOURCE PWR ON .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CHAN PWR [COUPLED] allows you to couple or uncouple channel power .

N

PORT POWER allows you to couple or uncouple port power .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Understanding the P ower Ranges

The built-in synthesized source contains a programmable step attenuator that allows you to

directly and accurately set power levels in eight di erent power ranges . Each range has a total

span of 25 dB . The eight ranges cover the instrument's full operating range from +10 dBm to

85 dBm (see Figure 6-6). A power range can be selected either manually or automatically .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PWR RANGE AUTO , you can enter any power level within the total operating range

of the instrument and the source attenuator will automatically switch to the corresponding

range .

Each range overlaps its adjacent ranges by 15 dB , therefore , certain power levels are

designated to cause the attenuator to switch to the next range so that optimum (leveled)

performance is maintained. These transition points exist at 10 dB from the top of a range and

at +5 dB from the bottom of a range . This leaves 10 dB of operating range . By turning the

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PORT POWER being the active function, you can hear the attenuator switch as

these transitions occur (see Figure 6-6).

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PWR RANGE MAN , you must rst enter the power ranges menu and manually select

the power range that corresponds to the power level you want to use . This is accomplished by

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

POWER RANGES softkey and then selecting one of the twelve available ranges . In

this mode , you will not be able to use the step keys , front panel knob , or keypad entry to select

power levels outside the range limits . This feature is necessary to maintain accuracy once a

measurement calibration is activated.

When a calibration is active , the power range selection is switched from auto to manual mode ,

and PRm appears on the display .

Application and Operation Concepts 6-13

Show 6 more labels / short text on this page
A utomatic mode · If you select · RPG knob with · Manual mode · If you select · pressing the
Page 281 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 281 · click to enlarge
Page 281

After measurement calibration, you can change the power within a range

and still maintain nearly full accuracy . In some cases better accuracy can

be achieved by changing the power within a range . It can be useful to set

di erent power levels for calibration and measurement to minimize the e ects

of sampler compression or noise oor .

If you decide to switch power ranges , the calibration is no longer valid and

accuracy is no longer speci ed. However , the analyzer leaves the correction on

even though it's invalid.

The annotation C? will be displayed whenever you change the power after

calibration.

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 281)
Fig. 122 · Chapter 6 Application and Operation Concepts

Figure 6-6. P ower Range Transitions in the A utomatic Mode

6-14 Application and Operation Concepts

Page 282 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 282 · click to enlarge
Page 282

P ower Coupling Options

There are two methods you can use to couple and uncouple power levels with the HP 8753D:

By uncoupling the channel powers , you e ectively have two separate sources . Uncoupling the

test ports allows you to have di erent power levels on each port.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CHAN POWER [COUPLED] toggles between coupled and uncoupled channel power . With the

channel power coupled, the power levels are the same on each channel. With the channel

power uncoupled, you can set di erent power levels for each channel. F or the channel

power to be uncoupled, the other channel stimulus functions must also be uncoupled

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

COUPLED CH OFF ).

T est port coupling

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PORT POWER [COUPLED] toggles between coupled and uncoupled test ports . With the test

ports coupled, the power level is the same at each port. With the ports uncoupled, you can

set a di erent power level at each port. This can be useful, for example , if you want to

simultaneously perform a gain and reverse isolation measurement on a high-gain ampli er using

the dual channel mode to display the results . In this case , you would want the power in the

) much lower than the power in the reverse direction (S

).

Application and Operation Concepts 6-15

Show 5 more labels / short text on this page
channel coupling · port coupling · Channel coupling · ( · forward direction (S
Page 283 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 283 · click to enlarge
Page 283

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SWEEP TIME softkey selects sweep time as the active entry and shows whether the

automatic or manual mode is active . The following explains the di erence between automatic

and manual sweep time:

Manual sweep time . As long as the selected sweep speed is within the capability of the

instrument, it will remain xed, regardless of changes to other measurement parameters . If

you change measurement parameters such that the instrument can no longer maintain the

selected sweep time , the analyzer will change to the fastest sweep time possible .

A uto sweep time . A uto sweep time continuously maintains the fastest sweep speed possible

with the selected measurement parameters .

Sweep time refers only to the time that the instrument is sweeping and taking data, and

does not include the time required for internal processing of the data, retrace time , or

bandswitching time . A sweep speed indicator " is displayed on the trace for sweep times longer

than 1.0 second. F or sweep times faster than 1.0 second the " indicator appears in the status

Manual Sweep Time Mode

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

When this mode is active , the softkey label reads

SWEEP TIME [MANUAL] . This mode is

engaged whenever you enter a sweep time greater than zero . This mode allows you to select a

xed sweep time . If you change the measurement parameters such that the current sweep time

is no longer possible , the analyzer will automatically increase to the next fastest sweep time

possible . If the measurement parameters are changed such that a faster sweep time is possible ,

the analyzer will not alter the sweep time while in this mode .

A uto Sweep Time Mode

When this mode is active , the softkey label reads

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SWEEP TIME [AUTO] . This mode is engaged

whenever you enter

as a sweep time . A uto sweep time continuously maintains the

fastest sweep time possible with the selected measurement parameters .

The minimum sweep time is dependent on the following measurement parameters:

the number of points selected

sweep-to-sweep averaging in dual channel display mode

In addition to the parameters listed above , the actual cycle time of the analyzer is also

dependent on the following measurement parameters:

6-16 Application and Operation Concepts

Show 11 more labels / short text on this page
Sweep Time · The · 0 · Minimum Sweep Time · IF bandwidth · error-correction · type of sweep · smoothing · limit test · trace math · marker statistics
Page 284 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 284 · click to enlarge
Page 284

time domain (Option 010 Only)

Use T able 6-1 to determine the minimum cycle time for the listed measurement parameters .

The values listed represent the minimum time required for a CW time measurement with

averaging o .

T able 6-1. Minimum Cycle Time (in seconds)

3700 Hz 3000 Hz 1000 Hz 300 Hz 100 Hz 30 Hz 10 Hz

11 0.0041 s 0.0055 s 0.012 s 0.037 s 0.108 s 0.359 s 1.14 s

51 0.0191 s 0.0255 s 0.060 s 0.172 s 0.504 s 1.660 s 5.30 s

101 0.0379 s 0.0505 s 0.120 s 0.341 s 0.998 s 3.300 s 10.5 s

201 0.0754 s 0.1005 s 0.239 s 0.679 s 1.990 s 6.600 s 20.9 s

401 0.1504 s 0.2005 s 0.476 s 1.355 s 3.960 s 13.10 s 41.7 s

801 0.3004 s 0.4005 s 0.951 s 2.701 s 7.910 s 26.10 s 83.3 s

1601 0.6004 s 0.8005 s 1.901 s 5.411 s 15.80 s 52.20 s 166.5 s

Application and Operation Concepts 6-17

Number of · IF Bandwidth · P oints

Page 285 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 285 · click to enlarge
Page 285

The trigger menu is used to select the type and number of groups for the sweep trigger . The

following is a description of the softkeys located within this menu:

NNNNNNNNNNNNNN

HOLD freezes the data trace on the display , and the analyzer stops sweeping and taking data.

The notation \Hld" is displayed at the left of the graticule . If the " indicator is on at the left

NNNNNNNNNNNNNNNNNNNN

side of the display , trigger a new sweep with

SINGLE .

NNNNNNNNNNNNNNNNNNNN

SINGLE takes one sweep of data and returns to the hold mode .

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NUMBER of GROUPS triggers a user-speci ed number of sweeps , and returns to the hold

mode . This function can be used to override the test set hold mode (indicated by the

switch (Option 007) and attenuator are not protected against unwanted continuous switching

. This occurs in a full two-port calibration, in a measurement of two di erent parameters

that require power out from both ports , or when the channels are uncoupled and a di erent

power level is set for each channel.

If averaging is on, the number of groups should be at least equal to the averaging factor

selected, to allow measurement of a fully averaged trace . Entering a number of groups resets

the averaging counter to 1.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CONTINUOUS is the standard sweep mode . The sweep is triggered automatically and

continuously and the trace is updated with each sweep .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRIGGER: TRIG OFF switches o external trigger mode .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EXT TRIG ON SWEEP is used when the sweep is triggered on an externally generated signal

that is connected to the rear panel EXT TRIGGER input. The sweep is started with a high

to low transition of a TTL signal. If this key is pressed when no external trigger signal is

connected, the notation ''Ext'' is displayed at the left side of the display to indicate that

the analyzer is waiting for a trigger . When a trigger signal is connected, the ''Ext'' notation

is replaced by the sweep speed indicator either in the status notations area or on the trace .

External trigger mode is allowed in every sweep mode .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EXT TRIG ON POINT is similar to the trigger on sweep , but triggers on each data point in a

sweep .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MANUAL TRG ON POINT waits for a manual trigger for each point. Subsequent pressing of

this softkey triggers each measurement. The annotation ''man'' appears at the left side of

the display when the instrument is waiting for the trigger to occur . This feature is useful in a

test sequence when an external device or instrument requires changes at each point.

6-18 Application and Operation Concepts

Trigger Menu

Page 286 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 286 · click to enlarge
Page 286

Source Attenuator Switch Protection

The programmable step attenuator of the source can be switched between port 1 and port 2

when the test port power is uncoupled or between channel 1 and channel 2 when the channel

power is uncoupled. T o avoid premature wear of the attenuator , measurement con gurations

requiring continuous switching between di erent power ranges are not allowed.

F or example , channels 1 and 2 of the analyzer are decoupled, power levels in two di erent

ranges are selected for each channel, and dual channel display is engaged. T o prevent

continuous switching between the two power ranges , the analyzer automatically engages

the test set hold mode after measuring both channels once . The active channel continues

to be updated each sweep while the inactive channel is placed in the hold mode . (The

status annotation tsH appears on the left side of the display .) If averaging is on, the test

set hold mode does not engage until the speci ed number of sweeps is completed. The

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NUMBER OF GROUPS (see \Trigger Menu") softkeys can override this

protection feature .

Allowing Repetitive Switching of the Attenuator

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NUMBER OF GROUPS (see \Trigger Menu") softkeys allow

measurements which demand repetitive switching of the step attenuator . Use these softkeys

with caution; repetitive switching can cause premature wearing of the attenuator .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MEASURE RESTART causes one measurement to occur before activating the test set hold

mode .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NUMBER OF GROUPS (see \Trigger Menu") causes a speci ed number of measurements to

occur before activating the test set hold mode .

Application and Operation Concepts 6-19

MEASURE RESTART and · The · MEASURE RESTART and

Page 287 of the Agilent Technologies 8753D user manual — COUPLED CH OFF
Page 287 · click to enlarge
Page 287

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

COUPLED CH on OFF toggles the channel coupling of stimulus values . With

COUPLED CH ON

(the preset condition), both channels have the same stimulus values . (The inactive channel

takes on the stimulus values of the active channel.)

In the stimulus coupled mode , the following parameters are coupled:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Coupling of stimulus values for the two channels is independent of

DUAL CHAN on OFF in

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

the display menu and

MARKERS: UNCOUPLED in the marker mode menu.

COUPLED CH OFF

activates an alternate sweep function when dual channel display is on. In this mode the

analyzer alternates between the two sets of stimulus values and displays the measurement data

of both channels .

6-20 Application and Operation Concepts

Show 11 more labels / short text on this page
Channel Stimulus Coupling · frequency · number of points · source power · number of groups · IF bandwidth · sweep time · trigger type · gating parameters · sweep type · power meter calibration
Page 288 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNN
Page 288 · click to enlarge
Page 288

The following softkeys are located within the sweep type menu. Among them are the ve

sweep types available .

NNNNNNNNNNNNNNNNNNNNNNNNNN

LIN FREQ (linear frequency sweep).

NNNNNNNNNNNNNNNNNNNNNNNNNN

LOG FREQ (logarithmic frequency sweep).

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LIST FREQ (list frequency sweep) provides access to the single/all segment menu.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

POWER SWEEP .

NNNNNNNNNNNNNNNNNNNNNNN

CW TIME (CW time sweep).

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EDIT LIST allows list frequencies to be entered or modi ed using the edit list menu and edit

subsweep menu.

The following sweep types will function with the interpolated error-correction feature

(described later):

The following sweep types will not function with the interpolated error correction feature

(described later):

Linear Frequency Sweep (Hz)

NNNNNNNNNNNNNNNNNNNNNNNNNN

LIN FREQ softkey activates a linear frequency sweep that is displayed on a standard

graticule with ten equal horizontal divisions . This is the preset default sweep type .

F or a linear sweep , sweep time is combined with the channel's frequency span to compute a

source sweep rate:

sweep rate = (frequency span) / (sweep time)

Since the sweep time may be a ected by various factors , the equation provided here is merely

an indication of the ideal (fastest) sweep rate . If the user-speci ed sweep time is greater

than 15 ms times the number of points , the sweep changes from a continuous ramp sweep

to a stepped CW sweep . Also , for 10 Hz or 30 Hz IF bandwidths the sweep is automatically

converted to a stepped CW sweep .

In the linear frequency sweep mode it is possible , with Option 010, to transform the data for

time domain measurements using the inverse F ourier transform technique .

Application and Operation Concepts 6-21

Show 7 more labels / short text on this page
Sweep Type Menu · linear frequency · power sweep · CW time · logarithmic frequency sweep · list frequency sweep · The
Page 289 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 289 · click to enlarge
Page 289

Logarithmic Frequency Sweep (Hz)

NNNNNNNNNNNNNNNNNNNNNNNNNN

LOG FREQ softkey activates a logarithmic frequency sweep mode . The source is stepped

in logarithmic increments and the data is displayed on a logarithmic graticule . This is slower

than a continuous sweep with the same number of points , and the entered sweep time may

therefore be changed automatically . F or frequency spans of less than two octaves , the sweep

type automatically reverts to linear sweep .

List Frequency Sweep (Hz)

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LIST FREQ softkey provides a user-de nable arbitrary frequency list mode . This list is

de ned and modi ed using the edit list menu and the edit subsweep menu. Up to 30 frequency

subsweeps (called \segments") of several di erent types can be speci ed, for a maximum total

of 1632 points . One list is common to both channels . Once a frequency list has been de ned

and a measurement calibration performed on the full frequency list, one or all of the frequency

segments can be measured and displayed without loss of calibration.

N

LIST FREQ key is pressed, the network analyzer sorts all the de ned frequency

NNNNNNNNNNNNNNNNNNNNNNNNNNNN

segments into CW points in order of increasing frequency . It then measures each point and

displays a single trace that is a composite of all data taken. If duplicate frequencies exist, the

analyzer makes multiple measurements on identical points to maintain the speci ed number of

points for each subsweep . Since the frequency points may not be distributed evenly across

the display , the display resolution may be uneven, and more compressed in some parts of the

trace than in others . However , the stimulus and response readings of the markers are always

accurate . Because the list frequency sweep is a stepped CW sweep , the sweep time is slower

than for a continuous sweep with the same number of points .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LIST FREQ softkey provides access to the segment menu, which allows you to select

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

any single segment (

SINGLE SEG SWEEP ) in the frequency list or all of the segments

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ALL SEGS SWEEP ) in the frequency list.

Refer to \Modifying List Frequencies" in this section for information on how to enter or modify

the list frequencies . If no list has been entered, the message CAUTION: LIST TABLE EMPTY is

displayed.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

A tabular printout of the frequency list data can be obtained using the

in the copy menu.

6-22 Application and Operation Concepts

Show 7 more labels / short text on this page
The · The · When the · Segment Menu · The · ( · LIST VALUES function
Page 290 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 290 · click to enlarge
Page 290

P ower Sweep (dBm)

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

POWER SWEEP softkey turns on a power sweep mode that is used to characterize

power-sensitive circuits . In this mode , power is swept at a single frequency , from a start power

value to a stop power value , selected using the

keys and the entry block.

ST ART

STOP

This feature is convenient for such measurements as gain compression or A GC (automatic gain

N

NNNNNNNNNNNNNNNNNNNNNN

control) slope . T o set the frequency of the power sweep , use

CW FREQ in the stimulus menu.

The span of the swept power is limited to being equal to or within one of the eight pre-de ned

power ranges . The attenuator will not switch to a di erent power range while in the power

sweep mode . Therefore , when performing a power sweep , power range selection will

automatically switch to the manual mode .

In power sweep , the entered sweep time may be automatically changed if it is less than the

minimum required for the current con guration (number of points , IF bandwidth, averaging,

etc .).

CW Time Sweep (Seconds)

NNNNNNNNNNNNNNNNNNNNNNN

CW TIME softkey turns on a sweep mode similar to an oscilloscope . The analyzer is set to a

single frequency , and the data is displayed versus time . The frequency of the CW time sweep is

NNNNNNNNNNNNNNNNNNNNNNN

CW FREQ in the stimulus menu. In this sweep mode , the data is continuously sampled

at precise , uniform time intervals determined by the sweep time and the number of points

minus 1. The entered sweep time may be automatically changed if it is less than the minimum

required for the current instrument con guration.

In time domain using Option 010, the CW time mode data is translated to frequency domain,

and the x-axis becomes frequency . In this mode , the instrument can be used as a spectrum

analyzer to measure signal purity , or for low frequency (

modulation signals .

In addition to the previous sweep types , there are also two di erent sweep modes . These

can be accessed through the correction menu by pressing

NNNNNNNNNNNNNN

MORE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ALTERNATE A and B

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CHOP A and B . Refer to \Alternate and Chop Sweep Modes" in the \Measurement

Calibration" section.

List frequencies can be entered or modi ed using the edit list and edit subsweep menus .

Application of the functions in these menus is described below .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EDIT LIST softkey within the sweep type menu provides access to the edit list menu.

This menu is used to edit the list of frequency segments (subsweeps) de ned with the edit

subsweep menu, described next. Up to 30 frequency subsweeps can be speci ed, for a

maximum of 1632 points . The segments do not have to be entered in any particular order:

the analyzer automatically sorts them and shows them on the display in increasing order of

start frequency . This menu determines which entry on the list is to be modi ed, while the edit

subsweep menu is used to make changes in the frequency or number of points of the selected

entry .

Application and Operation Concepts 6-23

Show 10 more labels / short text on this page
The · and · The · set with · Selecting Sweep Modes · ] · or · Modifying List Frequencies · Edit list menu · The
Page 291 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 291 · click to enlarge
Page 291

NNNNNNNNNNNNNN

NNNNNNNNNNN

ADD softkey within the edit list menu will display the edit subsweep menu.

This menu lets you select measurement frequencies arbitrarily . Using this menu it is possible to

de ne the exact frequencies to be measured on a point-by-point basis . F or example the sweep

could include 100 points in a narrow passband, 100 points across a broad stop band, and 50

points across the third harmonic response . The total sweep is de ned with a list of subsweeps .

The frequency subsweeps , or segments , can be de ned in any of the following terms:

The subsweeps can overlap , and do not have to be entered in any particular order . The

analyzer sorts the segments automatically and lists them on the display in order of increasing

start frequency , even if they are entered in center/span format. If duplicate frequencies

exist, the analyzer makes multiple measurements on identical points to maintain the speci ed

number of points for each subsweep . The data is shown on the display as a single trace that is

a composite of all data taken. The trace may appear uneven because of the distribution of the

data points , but the frequency scale is linear across the total range .

Once the list frequencies have been de ned or modi ed, the list frequency sweep mode can be

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LIST FREQ softkey in the sweep type menu (see \List Frequency Sweep").

The frequency list parameters can also be saved with an instrument state .

6-24 Application and Operation Concepts

Show 9 more labels / short text on this page
Edit subsweep menu · Using the · EDIT or · start/stop/number of points · start/stop/step · center/span/number of points · center/span/step · CW frequency · selected with the
Page 292 of the Agilent Technologies 8753D user manual — MARKER
Page 292 · click to enlarge
Page 292
Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 292)
Fig. 123 · Chapter 6 Application and Operation Concepts

Figure 6-7. Response Function Block

The following response function block keys are used to de ne and control the following

functions of the active channel.

MEAS

measurement parameters

F ORMA T

data format

SCALE REF

DISPLA Y

display functions

A VG

noise reduction alternatives

CAL

]calibration functions

MARKER

MARKER F CTN

display markers

The current values for the major response functions of the active channel are displayed in

speci c locations along the top of the display . In addition, certain functions accessed through

the keys in this block are annotated in the status notations area at the left side of the display .

An illustration of the analyzer's display showing the locations of these information labels is

provided in Chapter 1, \HP 8753D Description and Options ."

Application and Operation Concepts 6-25

Show 9 more labels / short text on this page
Response Functions · · · · · · · ·
Page 293 of the Agilent Technologies 8753D user manual — MEAS
Page 293 · click to enlarge
Page 293

key provides access to the S-parameter menu which contains softkeys that can be

MEAS

used to select the parameters or inputs that de ne the type of measurement being performed.

S-parameters (scattering parameters) are a convention used to characterize the way a device

modi es signal ow . A brief explanation of the S-parameters of a two-port device is provided

here . F or additional details refer to Hewlett-P ackard Application Notes A/N 95-1 and A/N 154.

S-parameters are always a ratio of two complex (magnitude and phase) quantities . S-parameter

where the rst number (out) refers to the test-device port where the signal is emerging and

the second number (in) is the test-device port where the signal is incident. F or example , the

identi es the measurement as the complex ratio of the signal emerging at the

test device's port 2 to the signal incident at the test device's port 1.

Figure 6-8 is a representation of the S-parameters of a two-port device , together with an

equivalent owgraph. In the illustration, \a" represents the signal entering the device and \b"

represents the signal emerging. Note that a and b are not related to the A and B input ports on

the analyzer .

Agilent Technologies 8753D — MEAS diagram (page 293)
Fig. 124

Figure 6-8. S-P arameters of a Two-P ort Device

6-26 Application and Operation Concepts

Show 5 more labels / short text on this page
S-P arameters · The · Understanding S-P arameters · S out in · S-parameter S
Page 294 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 294 · click to enlarge
Page 294

S-parameters are exactly equivalent to the more common description terms below , requiring

only that the measurements be taken with all test device ports properly terminated.

S-P arameter De nition T est Set

S

= 0 Input re ection coecient FWD

S

= 0 F orward gain FWD

S

= 0 Reverse gain REV

S

= 0 Output re ection coecient REV

The S-P arameter Menu

The S-parameter menu allows you to de ne the input ports and test set direction for

S-parameter measurements . The analyzer automatically switches the direction of the

measurement according to the selections you made in this menu. Therefore , the analyzer can

measure all four S-parameters with a single connection. The S-parameter being measured is

labeled at the top left corner of the display .

The S-parameter menu contains the following softkeys:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ReflFWD S11 (A/R)

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TransFWD S21 (B/R)

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TransREV S12 (A/R)

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ReflREV S22 (B/R)

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ANALOG IN Aux Input

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CONVERSION [ ] provides access to the conversion menu.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INPUT PORTS provides access to the input ports menu.

This menu allows you to monitor voltage and frequency nodes , using the analog bus and

internal counter . F or more information, refer to Chapter 10, \Service Key Menus and Error

Messages" in the HP 8753D Network Analyzer Service Guide .

This menu converts the measured re ection or transmission data to the equivalent complex

impedance (Z) or admittance (Y) values . This is not the same as a two-port Y or Z parameter

conversion, as only the measured parameter is used in the equations . Two simple one-port

conversions are available , depending on the measurement con guration.

trace measured as re ection can be converted to equivalent parallel impedance or

admittance using the model and equations shown in Figure 6-9.

Application and Operation Concepts 6-27

Show 18 more labels / short text on this page
Direction · Description · b · a · a · b · a · a · b · a · a · b · a · a · Analog In Menu · Conversion Menu · An S · or S
Page 295 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 295 · click to enlarge
Page 295
Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 295)
Fig. 125 · Chapter 6 Application and Operation Concepts

Figure 6-9. Re ection Impedance and A dmittance Conversions

In a transmission measurement, the data can be converted to its equivalent series impedance or

admittance using the model and equations shown in Figure 6-10.

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 295)
Fig. 126 · Chapter 6 Application and Operation Concepts

Figure 6-10. Transmission Impedance and A dmittance Conversions

A void the use of Smith chart, SWR, and delay formats for display of Z and Y

conversions , as these formats are not easily interpreted.

Input P orts Menu

This menu allows you to de ne the input ports for power ratio measurements , or a single input

for magnitude only measurements of absolute power . Y ou cannot use single inputs for phase or

group delay measurements , or any measurements with averaging activated.

6-28 Application and Operation Concepts

Page 296 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNN
Page 296 · click to enlarge
Page 296

The F ormat Menu

key provides access to the format menu. This menu allows you to select the

F ORMA T

appropriate display format for the measured data. The following list identi es which formats

are available by means of which softkeys:

N

NNNNNNNNNNNNNNNNNNNNNN

LOG MAG

NNNNNNNNNNNNNNNNN

PHASE

NNNNNNNNNNNNNNNNN

DELAY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SMITH CHART

NNNNNNNNNNNNNNNNN

POLAR

NNNNNNNNNNNNNNNNNNNNNNN

LIN MAG

NNNNNNNNNNN

SWR

NNNNNNNNNNNNNN

REAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

IMAGINARY

The analyzer automatically changes the units of measurement to correspond with the displayed

format. Special marker menus are available for the polar and Smith formats , each providing

several di erent marker types for readout of values .

The selected display format of a particular S-parameter or input is assigned to that parameter .

Thus if di erent S-parameters are measured, even if only one channel is used, each parameter

is shown in its selected format each time it is displayed.

The illustrations below show a re ection measurement of a bandpass lter displayed in each of

the available formats .

Log Magnitude F ormat

NNNNNNNNNNNNNNNNNNNNNNN

LOG MAG softkey displays the log magnitude format. This is the standard Cartesian format

used to display magnitude-only measurements of insertion loss , return loss , or absolute power

in dB versus frequency . Figure 6-11 illustrates the bandpass lter re ection data in a log

magnitude format.

Application and Operation Concepts 6-29

The · The

Page 297 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNN
Page 297 · click to enlarge
Page 297
Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 297)
Fig. 127 · Chapter 6 Application and Operation Concepts

Figure 6-11. Log Magnitude F ormat

NNNNNNNNNNNNNNNNN

PHASE softkey displays a Cartesian format of the phase portion of the data, measured

in degrees . This format displays the phase shift versus frequency . Figure 6-12 illustrates the

phase response of the same lter in a phase-only format.

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNN diagram (page 297)
Fig. 128

Figure 6-12. Phase F ormat

Group Delay F ormat

NNNNNNNNNNNNNNNNN

DELAY softkey selects the group delay format, with marker values given in seconds .

Figure 6-13 shows the bandpass lter response formatted as group delay . Group delay

principles are described in the next few pages .

6-30 Application and Operation Concepts

Phase F ormat · The · The

Page 298 of the Agilent Technologies 8753D user manual — MARKER F CTN
Page 298 · click to enlarge
Page 298
Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 298)
Fig. 129 · Chapter 6 Application and Operation Concepts

Figure 6-13. Group Delay F ormat

Smith Chart F ormat

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SMITH CHART softkey displays a Smith chart format (see Figure 6-14). This is used

in re ection measurements to provide a readout of the data in terms of impedance . The

intersecting dotted lines on the Smith chart represent constant resistance and constant

reactance values , normalized to the characteristic impedance , Z

, of the system. Reactance

values in the upper half of the Smith chart circle are positive (inductive) reactance , and

those in the lower half of the circle are negative (capacitive) reactance . The default marker

readout is in ohms ( ) to measure resistance and reactance (R+jX). A dditional marker types are

available in the Smith marker menu.

The Smith chart is most easily understood with a full scale value of 1.0. If the scale per

division is less than 0.2, the format switches automatically to polar .

If the characteristic impedance of the system is not 50 ohms , modify the impedance value

recognized by the analyzer by pressing

NNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNN

SET Z0 (the impedance value)

.

CAL

MORE

An inverted Smith chart format for admittance measurements (Figure 6-14) is also available .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

A ccess this by selecting

SMITH CHART in the format menu, and pressing

MARKER F CTN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNN

MKR MODE MENU

SMITH MKR MENU

G+jB MKR . The Smith chart is inverted and marker values

are read out in siemens (S) to measure conductance and susceptance (G+jB).

Application and Operation Concepts 6-31

The · ·

Page 299 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNN
Page 299 · click to enlarge
Page 299
Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 299)
Fig. 130 · Chapter 6 Application and Operation Concepts

Figure 6-14. Standard and Inverse Smith Chart F ormats

P olar F ormat

NNNNNNNNNNNNNNNNN

POLAR softkey displays a polar format (see Figure 6-15). Each point on the polar format

corresponds to a particular value of both magnitude and phase . Quantities are read vectorally:

the magnitude at any point is determined by its displacement from the center (which has zero

value), and the phase by the angle counterclockwise from the positive x-axis . Magnitude is

scaled in a linear fashion, with the value of the outer circle usually set to a ratio value of 1.

Since there is no frequency axis , frequency information is read from the markers .

The default marker readout for the polar format is in linear magnitude and phase . A log

magnitude marker and a real/imaginary marker are available in the polar marker menu.

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNN diagram (page 299)
Fig. 131

Figure 6-15. P olar F ormat

6-32 Application and Operation Concepts

The

Page 300 of the Agilent Technologies 8753D user manual — NNNNNNNNNNN
Page 300 · click to enlarge
Page 300

Linear Magnitude F ormat

NNNNNNNNNNNNNNNNNNNNNNN

LIN MAG softkey displays the linear magnitude format (see Figure 6-16). This is a

Cartesian format used for unitless measurements such as re ection coecient magnitude or

transmission coecient magnitude , and for linear measurement units . It is used for display of

conversion parameters and time domain transform data.

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNNNNN diagram (page 300)
Fig. 132

Figure 6-16. Linear Magnitude F ormat

NNNNNNNNNNN

SWR softkey reformats a re ection measurement into its equivalent SWR (standing wave

ratio) value (see Figure 6-17 ). SWR is equivalent to (1 + )/(1 ), where is the re ection

coecient. Note that the results are valid only for re ection measurements . If the SWR format

is used for measurements of S

the results are not valid.

Agilent Technologies 8753D — NNNNNNNNNNN diagram (page 300)
Fig. 133

Figure 6-17. Typical SWR Display

Application and Operation Concepts 6-33

Show 4 more labels / short text on this page
The · SWR F ormat · The · or S
Page 301 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 301 · click to enlarge
Page 301

NNNNNNNNNNNNNN

REAL softkey displays only the real (resistive) portion of the measured data on a Cartesian

format (see Figure 6-18). This is similar to the linear magnitude format, but can show both

positive and negative values . It is primarily used for analyzing responses in the time domain,

and also to display an auxiliary input voltage signal for service purposes .

Agilent Technologies 8753D — NNNNNNNNNNNNNN diagram (page 301)
Fig. 134

Figure 6-18. Real F ormat

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

IMAGINARY softkey displays only the imaginary (reactive) portion of the measured data

on a Cartesian format. This format is similar to the real format except that reactance data is

displayed on the trace instead of impedance data.

6-34 Application and Operation Concepts

Show 4 more labels / short text on this page
Real F ormat · The · Imaginary F ormat · The
Page 302 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 302 · click to enlarge
Page 302

F or many networks , the amount of insertion phase is not as important as the linearity of the

phase shift over a range of frequencies . The analyzer can measure this linearity and express

it in two di erent ways: directly , as deviation from linear phase , or as group delay , a derived

value .

Group delay is the measurement of signal transmission time through a test device . It is de ned

as the derivative of the phase characteristic with respect to frequency . Since the derivative

is basically the instantaneous slope (or rate of change of phase with respect to frequency), a

perfectly linear phase shift results in a constant slope , and therefore a constant group delay

(see Figure 6-19).

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 302)
Fig. 135 · Chapter 6 Application and Operation Concepts

Figure 6-19. Constant Group Delay

order (deviations from linear) components . The linear component can be attributed to the

electrical length of the test device , and represents the average signal transit time . The higher

order components are interpreted as variations in transit time for di erent frequencies , and

represent a source of signal distortion (see Figure 6-20).

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 302)
Fig. 136 · Chapter 6 Application and Operation Concepts

Figure 6-20. Higher Order Phase Shift

The analyzer computes group delay from the phase slope . Phase data is used to nd the phase

change , , over a speci ed frequency aperture , f, to obtain an approximation for the rate of

change of phase with frequency (see Figure 6-21). This value ,

, represents the group delay in

seconds assuming linear phase change over f. It is important that be 180

, or errors will

Application and Operation Concepts 6-35

Group Delay Principles · g ·

Page 303 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 303 · click to enlarge
Page 303

result in the group delay data. These errors can be signi cant for long delay devices . Y ou can

verify that is 180

by increasing the number of points or narrowing the frequency span (or

both) until the group delay data no longer changes .

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 303)
Fig. 137 · Chapter 6 Application and Operation Concepts

Figure 6-21. Rate of Phase Change V ersus Frequency

When deviations from linear phase are present, changing the frequency step can result in

di erent values for group delay . Note that in this case the computed slope varies as the

aperture f is increased (see Figure 6-22 ). A wider aperture results in loss of the ne grain

variations in group delay . This loss of detail is the reason that in any comparison of group delay

data, it is important to know the aperture that was used to make the measurement.

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 303)
Fig. 138 · Chapter 6 Application and Operation Concepts

Figure 6-22. V ariations in Frequency Aperture

In determining the group delay aperture , there is a tradeo between resolution of ne detail

and the e ects of noise . Noise can be reduced by increasing the aperture , but this will tend to

smooth out the ne detail. More detail will become visible as the aperture is decreased, but the

noise will also increase , possibly to the point of obscuring the detail. A good practice is to use a

smaller aperture to assure that small variations are not missed, then increase the aperture to

smooth the trace .

6-36 Application and Operation Concepts

Page 304 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 304 · click to enlarge
Page 304

The default group delay aperture is the frequency span divided by the number of points across

the display . T o set the aperture to a di erent value , turn on smoothing in the average menu,

and vary the smoothing aperture . The aperture can be varied up to 20% of the span swept.

Group delay measurements can be made on linear frequency , log frequency , or list frequency

sweep types (not in CW or power sweep). Group delay aperture varies depending on the

frequency spacing and point density , therefore the aperture is not constant in log and list

frequency sweep modes . In list frequency mode , extra frequency points can be de ned to

ensure the desired aperture .

T o obtain a readout of aperture values at di erent points on the trace , turn on a marker . Then

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

A VG

SMOOTHING APERTURE . Smoothing aperture becomes the active function, and as the

aperture is varied its value in Hz is displayed below the active entry area.

Application and Operation Concepts 6-37

press ·

Page 305 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 305 · click to enlarge
Page 305

key provides access to the scale reference menu. Softkeys within this menu

SCALE REF

can be used to de ne the scale in which measured data is to be displayed, as well as simulate

phase o set and electrical delay . The following softkeys are located within the scale reference

menu.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AUTO SCALE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SCALE/DIV

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REFERENCE POSITION

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REFERENCE VALUE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER ! REFERENCE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ELECTRICAL DELAY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PHASE OFFSET

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

COAXIAL DELAY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

WAVEGUIDE DELAY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ELECTRICAL DELAY softkey adjusts the electrical delay to balance the phase of the test

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

device . This softkey must be used in conjunction with

WAVEGUIDE DELAY

(with cut-o frequency) in order to identify which type of transmission line the delay is being

added to .

Electrical delay simulates a variable length lossless transmission line , which can be added to or

removed from a receiver input to compensate for interconnecting cables , etc . This function

is similar to the mechanical or analog \line stretchers" of other network analyzers . Delay is

annotated in units of time with secondary labeling in distance for the current velocity factor .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

With this feature , and with

MARKER ! DELAY (see \Using Markers"), an equivalent length of

air- lled, lossless transmission line is added or subtracted according to the following formula:

Leng th (meter s) =

(F r eq (M H z ) 1: 20083)

Once the linear portion of the test device's phase has been removed, the equivalent length of

the lossless , transmission line can be read out in the active marker area. If the average relative

) of the test device is known over the frequency span, the length calculation can

be adjusted to indicate the actual length of the test device more closely . This can be done by

entering the relative velocity factor for the test device using the calibrate more menu. The

relative velocity factor for a given dielectric can be calculated by:

V el ocity F actor =

assuming a relative permeability of 1.

6-38 Application and Operation Concepts

Show 10 more labels / short text on this page
Scale Reference Menu · The · Electrical Delay · The · COAXIAL DELAY or · permittivity (" · r · p · " · r
Page 306 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 306 · click to enlarge
Page 306

key provides access to the display menu, which controls the memory math

DISPLA Y

functions and leads to other menus associated with display functions .

The following softkeys are located within the display menu:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DUAL CHAN on OFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DISPLAY: DATA

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DISPLAY: MEMORY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DISPLAYDATA and MEMORY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DISPLAY: DATA/MEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DISPLAY: DATA - MEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DISPLAY: DATA ! MEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SPLIT DISP ON off

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

BEEP DONE ON off

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

BEEP WARN on OFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ADJUST DISPLAY

NNNNNNNNNNNNNNNNN

TITLE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

D2/D1 TO D2 on OFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FREQUENCY BLANK

The analyzer has two available memory traces , one per channel. Memory traces are totally

channel dependent: channel 1 cannot access the channel 2 memory trace or vice versa.

Memory traces can be saved with instrument states: one memory trace can be saved per

channel per saved instrument state . There are up to 31 save/recall registers available , so the

total number of memory traces that can be present is 64 including the two active for the

current instrument state . The memory data is stored as full precision, complex data.

Y ou may not be able to store 31 instrument states if they include a large

amount of calibration data. The calibration data contributes considerably to the

size of the instrument state le and therefore the available memory may be full

prior to lling all 31 registers .

Application and Operation Concepts 6-39

Display Menu · The

Page 307 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 307 · click to enlarge
Page 307

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DUAL CHAN on OFF softkey toggles between display of both measurement channels or

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

the active channel only . This is used in conjunction with

SPLIT DISP ON off in the display

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

more menu to display both channels . With

SPLIT DISP on OFF the two traces are overlaid

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

on a single graticule (see Figure 6-23a); with

SPLIT DISP ON off the measurement data

is displayed on two half-screen graticules one above the other (see Figure 6-23b). Current

parameters for the two displays are annotated separately .

The stimulus functions of the two channels can also be controlled independently using

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

COUPLED CH ON off in the stimulus menu. In addition, the markers can be controlled

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

independently for each channel using

MARKERS: UNCOUPLED in the marker mode menu.

Dual Channel Mode with Decoupled Channel P ower

By decoupling the channel power or port power and using the dual channel mode , you can

simultaneously view two measurements having di erent power levels . However , there are two

con gurations that may not appear to function \properly".

1. Channel 1 requires one attenuation value and channel 2 requires a di erent value . Since

one attenuator is used for both testports , this would cause the attenuator to continuously

switch power ranges .

2. With Option 007 (mechanical transfer switch), channel 1 is driving one test port and channel

2 is driving the other test port. This would cause the test port transfer switch to continually

cycle . The instrument will not allow the transfer switch or attenuator to continuously

switch ranges in order to update these measurements without the direct intervention of the

operator .

If one of the above conditions exists , the test set hold mode will engage , and the status

function in only one of the two measurement paths . T o update both measurement setups , press

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MENU

MEASURE RESTART . Refer to \Source Attenuator Switch Protection" earlier in this

chapter for more information on this condition.

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 307)
Fig. 139

Figure 6-23. Dual Channel Displays

6-40 Application and Operation Concepts

Show 4 more labels / short text on this page
Dual Channel Mode · The · ·
Page 308 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 308 · click to enlarge
Page 308

Two trace math operations are implemented:

NNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNN

DATA MEM (datamemory)

NNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNN

(Note that normalization is

DATA MEM .) Memory traces are saved and recalled

and trace math is done immediately after error-correction. This means that any data processing

done after error-correction, including parameter conversion, time domain transformation

(Option 010), scaling, etc ., can be performed on the memory trace . Y ou can also use trace math

as a simple means of error-correction, although that is not its main purpose .

All data processing operations that occur after trace math, except smoothing and gating, are

identical for the data trace and the memory trace . If smoothing or gating is on when a memory

trace is saved, this state is maintained regardless of the data trace smoothing or gating status .

If a memory trace is saved with gating or smoothing on, these features can be turned on or o

in the memory-only display mode .

The actual memory for storing a memory trace is allocated only as needed. The memory trace

is cleared on instrument preset, power on, or instrument state recall.

If sweep mode or sweep range is di erent between the data and memory traces , trace math is

allowed, and no warning message is displayed. If the number of points in the two traces is

di erent, the memory trace is not displayed nor rescaled. However , if the number of points for

the data trace is changed back to the number of points in the memory , the memory trace can

then be displayed.

If trace math or display memory is requested and no memory trace exists , the message

A djusting the Colors of the Display

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ADJUST DISPLAY softkey provides access to the adjust display menu. The following

softkeys are located within this menu:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INTENSITY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

BACKGROUND INTENSITY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MODIFY COLORS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DEFAULT COLORS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

BLANK DISPLAY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SAVE COLORS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECALL COLORS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

T o adjust the intensity of the display , press

INTENSITY and rotate the front panel knob , use

keys , or use the numerical keypad to set the intensity value between 50 and 100

percent. Lowering the intensity may prolong the life of the LCD .

Application and Operation Concepts 6-41

Show 8 more labels / short text on this page
Memory Math Functions · DATA/MEM (data/memory) · DATA/MEM not · The · Setting Display Intensity · the · * · +
Page 309 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNN
Page 309 · click to enlarge
Page 309

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

T o set all the display elements to the factory-de ned default colors , press

DEFAULT COLORS .

NNNNNNNNNNNNNNNNNNNN

PRESET does not reset or change colors to the default color values . However ,

cycling power to the instrument will reset the colors to the default color

values .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

BLANK DISPLAY switches o the analyzer display while leaving the instrument in

its current measurement state . This feature may be helpful in prolonging the life of the LCD

in applications where the analyzer is left unattended (such as in an automated test system).

Turning the front panel knob or pressing any front panel key will restore normal display

operation.

Saving Modi ed Colors

T o save a modi ed color set, press

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SAVE COLORS . Modi ed colors are not part of a saved

instrument state and are lost unless saved using these softkeys .

Recalling Modi ed Colors

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

T o recall the previously saved color set, press

RECALL COLORS .

The Modify Colors Menu

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MODIFY COLORS softkey within the adjust display menu provides access to the modify

colors menu.

The modify colors menu allows you to adjust the colors on your analyzer's display . The default

colors in this instrument have been scienti cally chosen to maximize your ability to discern the

di erence between the colors , and to comfortably and e ectively view the colors . These colors

are recommended for normal use because they will provide a suitable contrast that is easy to

view for long periods of time .

Y ou may choose to change the default colors to suit environmental needs , individual

preferences , or to accommodate color de cient vision. Y ou can use any of the available colors

for any of the seven display elements listed by the softkey names below:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CH1 DATA/LIMIT LN

NNNNNNNNNNNNNNNNNNNNNNN

CH1 MEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CH2 DATA/LIMIT LN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CH2 MEM/REF LINE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GRATICULE/TEXT

NNNNNNNNNNNNNNNNNNNNNNN

WARNING

NNNNNNNNNNNNNN

TEXT

T o change the color of a display elements , press the softkey for that element (such as

NNNNNNNNNNNNNNNNNNNNNNNNNN

CH1 DATA ). Then press

NNNNNNNNNNNNNN

TINT and turn the analyzer front panel knob , use the step keys or the

numeric keypad, until the desired color appears .

6-42 Application and Operation Concepts

Show 4 more labels / short text on this page
Setting Default Colors · Blanking the Display · Pressing · The
Page 310 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNN
Page 310 · click to enlarge
Page 310

Maximum viewing angle with the LCD display is achieved when primary colors

or a combination of them are selected at full brightness (100%). The following

table lists the recommended colors and their corresponding tint numbers .

T able 6-2. Display Colors with Maximum Viewing Angle

Display Color Tint Brightness Color

Red 0 100 100

Y ellow 17 100 100

Green 33 100 100

Cyan 50 100 100

Blue 67 100 100

Magenta 83 100 100

Color is comprised of three parameters:

Tint: The continuum of hues on the color wheel, ranging from red, through green and blue ,

and back to red.

Brightness: A measure of the brightness of the color .

Color: The degree of whiteness of the color . A scale from white to pure color .

The most frequently occurring color de ciency is the inability to distinguish red, yellow ,

and green from one another . Confusion between these colors can usually be eliminated by

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

increasing the brightness between the colors . T o accomplish this , press the

BRIGHTNESS

softkey and turn the analyzer front panel knob . If additional adjustment is needed, vary the

NNNNNNNNNNNNNNNNN

degree of whiteness of the color . T o accomplish this , press the

COLOR softkey and turn the

analyzer front panel knob .

Color changes and adjustments remain in e ect until changed again in these

menus or the analyzer is powered o and then on again. Cycling the power

changes all color adjustments to default values . Preset does not a ect color

selection.

Application and Operation Concepts 6-43

White 100 0

Page 311 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 311 · click to enlarge
Page 311

key is used to access three di erent noise reduction techniques: sweep-to-sweep

A VG

averaging, display smoothing, and variable IF bandwidth. All of these can be used

simultaneously . A veraging and smoothing can be set independently for each channel, and the

IF bandwidth can be set independently if the stimulus is uncoupled.

The following softkeys are located within the averaging menu:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AVERAGING RESTART

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AVERAGING FACTOR

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SMOOTHING APERTURE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

IF BW [ ]

A veraging computes each data point based on an exponential average of consecutive sweeps

weighted by a user-speci ed averaging factor . Each new sweep is averaged into the trace until

the total number of sweeps is equal to the averaging factor , for a fully averaged trace . Each

point on the trace is the vector sum of the current trace data and the data from the previous

sweep . A high averaging factor gives the best signal-to-noise ratio , but slows the trace update

time . Doubling the averaging factor reduces the noise by 3 dB . A veraging is used for ratioed

measurements: if it is attempted for a single-input measurement (e .g. A or B), the message

e ect of averaging on a log magnitude format trace .

If you switch power ranges with averaging on, the average will restart.

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 311)
Fig. 140

Figure 6-24. E ect of A veraging on a Trace

6-44 Application and Operation Concepts

Show 5 more labels / short text on this page
A veraging Menu · The · AVERAGING ON off · SMOOTHING ON off · A veraging
Page 312 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 312 · click to enlarge
Page 312

Smoothing (similar to video ltering) averages the formatted active channel data over a portion

of the displayed trace . Smoothing computes each displayed data point based on one sweep

only , using a moving average of several adjacent data points for the current sweep . The

smoothing aperture is a percent of the swept stimulus span, up to a maximum of 20%.

Rather than lowering the noise oor , smoothing nds the mid-value of the data. Use it

to reduce relatively small peak-to-peak noise values on broadband measured data. Use a

suciently high number of display points to avoid misleading results . Do not use smoothing for

measurements of high resonance devices or other devices with wide trace variations , as it will

introduce errors into the measurement.

Smoothing is used with Cartesian and polar display formats . It is also the primary way to

control the group delay aperture , given a xed frequency span. (Refer to \Group Delay

Principles" earlier in this section.) In polar display format, large phase shifts over the

smoothing aperture will cause shifts in amplitude , since a vector average is being computed.

Figure 6-25 illustrates the e ect of smoothing on a log magnitude format trace .

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 312)
Fig. 141 · Chapter 6 Application and Operation Concepts

Figure 6-25. E ect of Smoothing on a Trace

IF bandwidth reduction lowers the noise oor by digitally reducing the receiver input

bandwidth. It works in all ratio and non-ratio modes . It has an advantage over averaging as it

reliably lters out unwanted responses such as spurs , odd harmonics , higher frequency spectral

noise , and line-related noise . Sweep-to-sweep averaging, however , is better at ltering out very

low frequency noise . A tenfold reduction in IF bandwidth lowers the measurement noise oor

by about 10 dB . Bandwidths less than 300 Hz provide better harmonic rejection than higher

bandwidths .

Another di erence between sweep-to-sweep averaging and variable IF bandwidth is the sweep

time . A veraging displays the rst complete trace faster but takes several sweeps to reach a

fully averaged trace . IF bandwidth reduction lowers the noise oor in one sweep , but the

sweep time may be slower . Figure 6-26 illustrates the di erence in noise oor between a trace

measured with a 3000 Hz IF bandwidth and with a 10 Hz IF bandwidth.

Application and Operation Concepts 6-45

Smoothing · IF Bandwidth Reduction

Page 313 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 313 · click to enlarge
Page 313
Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 313)
Fig. 142 · Chapter 6 Application and Operation Concepts

Figure 6-26. IF Bandwidth Reduction

Another capability that can be used for e ective noise reduction is the marker

statistics function, which computes the average value of part or all of the

formatted trace .

If your instrument is equipped with Option 085 (High P ower System), another

way of increasing dynamic range is to increase the input power to the test

device using a booster ampli er .

6-46 Application and Operation Concepts

Page 314 of the Agilent Technologies 8753D user manual — MARKER
Page 314 · click to enlarge
Page 314

key displays a movable active marker on the screen and provides access to a

MARKER

series of menus to control up to ve display markers for each channel. Markers are used to

obtain numerical readings of measured values . They also provide capabilities for reducing

measurement time by changing stimulus parameters , searching the trace for speci c values , or

statistically analyzing part or all of the trace . Figure 6-27 illustrates the displayed trace with all

markers on and marker 2 the active marker .

Agilent Technologies 8753D — MARKER diagram (page 314)
Fig. 143

Figure 6-27. Markers on Trace

Markers have a stimulus value (the x-axis value in a Cartesian format) and a response value

(the y-axis value in a Cartesian format). In a polar or Smith chart format, the second part of a

complex data pair is also provided as an auxiliary response value . When a marker is activated

and no other function is active , its stimulus value is displayed in the active entry area and can

be controlled with the knob , the step keys , or the numeric keypad. The active marker can be

moved to any point on the trace , and its response and stimulus values are displayed at the top

right corner of the graticule for each displayed channel, in units appropriate to the display

format. The displayed marker response values are valid even when the measured data is above

or below the range displayed on the graticule .

Marker values are normally continuousthat is , they are interpolated between measured

points . Or , they can be set to read only discrete measured points . The markers for the two

channels normally have the same stimulus values , or they can be uncoupled so that each

channel has independent markers , regardless of whether stimulus values are coupled or dual

channel display is on.

If both data and memory are displayed, the marker values apply to the data trace . If only

memory is displayed, the marker values apply to the memory trace . In a memory math display

(data/memory or datamemory), the marker values apply to the trace resulting from the

memory math function.

Application and Operation Concepts 6-47

Markers · The

Page 315 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNN
Page 315 · click to enlarge
Page 315

With the use of a reference marker , a delta marker mode is available that displays both the

stimulus and response values of the active marker relative to the reference . Any of the ve

markers or a xed point can be designated as the delta reference marker . If the delta reference

is one of the ve markers , its stimulus value can be controlled by the user and its response

value is the value of the trace at that stimulus value . If the delta reference is a xed marker ,

both its stimulus value and its response value can be set arbitrarily anywhere in the display

area (not necessarily on the trace).

Markers can be used to search for the trace maximum or minimum point or any other point

on the trace . The ve markers can be used together to search for speci ed bandwidth cuto

points and calculate the bandwidth and Q values . Statistical analysis uses markers to provide a

readout of the mean, standard deviation, and peak-to-peak values of all or part of the trace .

Basic marker operations are available in the menus accessed from the

MARKER

key . The

marker search and statistical functions , together with the capability for quickly changing

stimulus parameters with markers , are provided in the menus accessed from the

MARKER F CTN

key .

MARKER

key provides access to the marker menu. This menu allows you to turn the

display markers on or o , to designate the active marker , and to gain access to the delta

marker menu and the xed marker menu.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MODE MENU softkey within the marker menu provides access to the delta mode menu.

The delta reference is shown on the display as a small triangle , smaller than the inactive

marker triangles . If one of the markers is the reference , the triangle appears next to the

marker number on the trace .

The marker values displayed in this mode are the stimulus and response values of the active

marker minus the reference marker . If the active marker is also designated as the reference

marker , the marker values are zero .

Fixed Marker Menu. The

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FIXED MKR POSTION softkey within the delta mode menu provides

access to the xed marker menu. This menu is used to set the position of a xed reference

marker , indicated on the display by a small triangle . Both the stimulus value and the

response value of the xed marker can be set arbitrarily anywhere in the display area, and

need not be on the trace . The units are determined by the display format, the sweep type , and

the marker type .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

There are two ways to turn on the xed marker . One way is with the

REF = FIXED MKR

softkey in the delta marker menu. The other is with the

NNNNNNNNNNNNNNNNNNNNNNNNNN

MKR ZERO function in the marker

menu, which puts a xed reference marker at the present active marker position and makes

the marker stimulus and response values at that position equal to zero .

The softkeys in this menu make the values of the xed marker the active function. The marker

readings in the top right corner of the graticule are the stimulus and response values of the

active marker minus the xed reference marker . Also displayed in the top right corner is the

The stimulus value , response value , and auxiliary response value (the second part of a complex

data pair) can be individually examined and changed. This allows active marker readings that

are relative in amplitude yet absolute in frequency , or any combination of relative/absolute

NNNNNNNNNNNNNNNNNNNNNNNNNN

readouts . F ollowing a

MKR ZERO operation, this menu can be used to reset any of the xed

marker values to absolute zero for absolute readings of the subsequent active marker values .

6-48 Application and Operation Concepts

Show 5 more labels / short text on this page
· Marker Menu · The · Delta Mode Menu · The
Page 316 of the Agilent Technologies 8753D user manual — F ORMA T
Page 316 · click to enlarge
Page 316

If the format is changed while a xed marker is on, the xed marker values become invalid.

F or example , if the value o set is set to 10 dB with a log magnitude format, and the format

is then changed to phase , the value o set becomes 10 degrees . However , in polar and Smith

chart formats , the speci ed values remain consistent between di erent marker types for those

formats . Thus an R+jX marker set on a Smith chart format will retain the equivalent values if

it is changed to any of the other Smith chart markers .

MARKER F CTN

key provides access to the marker function menu. This menu provides

softkeys that use markers to quickly modify certain measurement parameters without going

through the usual key sequence . In addition, it provides access to two additional menus used

for searching the trace and for statistical analysis .

NNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER ! functions change certain stimulus and response parameters to make them

equal to the current active marker value . Use the knob or the numeric keypad to move the

marker to the desired position on the trace , and press the appropriate softkey to set the

speci ed parameter to that trace value . When the values have been changed, the marker can

again be moved within the range of the new parameters .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MKR SEARCH [ ] softkey within the marker function menu provides access to the

marker search menu. This menu is used to search the trace for a speci c amplitude-related

point, and place the marker on that point. The capability of searching for a speci ed

bandwidth is also provided. Tracking is available for a continuous sweep-to-sweep search. If

there is no occurrence of a speci ed value or bandwidth, the message TARGET VALUE NOT FOUND

is displayed.

NNNNNNNNNNNNNNNNNNNN

T arget Menu. The

TARGET softkey within the marker search menu provides access to the

target menu. This menu lets you place the marker at a speci ed target response value on

the trace , and provides search right and search left options . If there is no occurrence of the

speci ed value , the message TARGET VALUE NOT FOUND is displayed.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MKR MODE MENU softkey within the marker function menu provides access to the marker

mode menu.This menu provides di erent marker modes and leads to the following two menus:

P olar Marker Menu. This menu is used only with a polar display format, selectable using the

F ORMA T

key . In a polar format, the magnitude at the center of the circle is zero and the outer

circle is the full scale value set in the scale reference menu. Phase is measured as the angle

at the positive x-axis . The analyzer automatically calculates di erent

mathematical forms of the marker magnitude and phase values , selected using the softkeys in

this menu. Marker frequency is displayed in addition to other values regardless of the selection

of marker type .

Smith Marker Menu. This menu is used only with a Smith chart format, selected from

the format menu. The analyzer automatically calculates di erent mathematical forms of

the marker magnitude and phase values , selected using the softkeys in this menu. Marker

frequency is displayed in addition to other values for all marker types .

Application and Operation Concepts 6-49

Show 10 more labels / short text on this page
Marker Function Menu · The · The · Marker Search Menu · The · Marker Mode Menu · The · · counterclockwise from 0 ·
Page 317 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 317 · click to enlarge
Page 317

Measurement calibration is an accuracy enhancement procedure that e ectively removes the

system errors that cause uncertainty in measuring a test device . It measures known standard

devices , and uses the results of these measurements to characterize the system.

This section discusses the following topics:

de nition of accuracy enhancement

causes of measurement errors

characterization of microwave systematic errors

e ectiveness of accuracy enhancement

correcting for measurement errors

ensuring a valid calibration

calibrating for non-insertable devices

What is A ccuracy Enhancement?

A perfect measurement system would have in nite dynamic range , isolation, and directivity

characteristics , no impedance mismatches in any part of the test setup , and at frequency

response . In any high frequency measurement there are measurement errors associated with

the system that contribute uncertainty to the results . P arts of the measurement setup such as

interconnecting cables and signal-separation devices (as well as the analyzer itself) all introduce

variations in magnitude and phase that can mask the actual performance of the test device .

V ector accuracy enhancement, also known as measurement calibration or error-correction,

provides the means to simulate a nearly perfect measurement system.

F or example , crosstalk due to the channel isolation characteristics of the analyzer can

contribute an error equal to the transmission signal of a high-loss test device . F or re ection

measurements , the primary limitation of dynamic range is the directivity of the test setup .

The measurement system cannot distinguish the true value of the signal re ected by the test

device from the signal arriving at the receiver input due to leakage in the system. F or both

transmission and re ection measurements , impedance mismatches within the test setup cause

measurement uncertainties that appear as ripples superimposed on the measured data.

Error-correction simulates an improved analyzer system. During the measurement calibration

process , the analyzer measures the magnitude and phase responses of known standard devices ,

and compares the measurement with actual device data. The analyzer uses the results to

characterize the system and e ectively remove the system errors from the measurement data

of a test device , using vector math capabilities internal to the network analyzer .

When you use a measurement calibration, the dynamic range and accuracy of the measurement

are limited only by system noise and stability , connector repeatability , and the accuracy to

which the characteristics of the calibration standards are known.

6-50 Application and Operation Concepts

Show 5 more labels / short text on this page
Measurement Calibration · calibration considerations · modifying calibration kits · TRL*/LRM* calibration · power meter calibration
Page 318 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 318 · click to enlarge
Page 318

What Causes Measurement Errors?

Network analysis measurement errors can be separated into systematic , random, and drift

errors .

Correctable systematic errors are the repeatable errors that the system can measure . These are

errors due to mismatch and leakage in the test setup , isolation between the reference and test

signal paths , and system frequency response .

The system cannot measure and correct for the non-repeatable random and drift errors . These

errors a ect both re ection and transmission measurements . Random errors are measurement

variations due to noise and connector repeatability . Drift errors include frequency drift,

temperature drift, and other physical changes in the test setup between calibration and

measurement.

The resulting measurement is the vector sum of the test device response plus all error terms .

The precise e ect of each error term depends upon its magnitude and phase relationship to the

actual test device response .

In most high frequency measurements the systematic errors are the most signi cant source of

measurement uncertainty . Since each of these errors can be characterized, their e ects can be

e ectively removed to obtain a corrected value for the test device response . F or the purpose of

vector accuracy enhancement these uncertainties are quanti ed as directivity , source match,

load match, isolation (crosstalk), and frequency response (tracking). Each of these systematic

errors is described below .

Random and drift errors cannot be precisely quanti ed, so they must be treated as producing a

cumulative uncertainty in the measured data.

Normally a device that can separate the reverse from the forward traveling waves (a

directional bridge or coupler) is used to detect the signal re ected from the test device . Ideally

the coupler would completely separate the incident and re ected signals , and only the re ected

signal would appear at the coupled output, as illustrated in Figure 6-28 a.

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 318)
Fig. 144 · Chapter 6 Application and Operation Concepts

Figure 6-28. Directivity

However , an actual coupler is not perfect, as illustrated in Figure 6-28b . A small amount of

the incident signal appears at the coupled output due to leakage as well as re ection from the

termination in the coupled arm. Also , re ections from the coupler output connector appear at

the coupled output, adding uncertainty to the signal re ected from the device . The gure of

merit for how well a coupler separates forward and reverse waves is directivity . The greater

the directivity of the device , the better the signal separation. System directivity is the vector

sum of all leakage signals appearing at the analyzer receiver input. The error contributed by

directivity is independent of the characteristics of the test device and it usually produces the

major ambiguity in measurements of low re ection devices .

Application and Operation Concepts 6-51

Directivity

Page 319 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 319 · click to enlarge
Page 319

Source match is de ned as the vector sum of signals appearing at the analyzer receiver input

due to the impedance mismatch at the test device looking back into the source , as well as to

adapter and cable mismatches and losses . In a re ection measurement, the source match error

signal is caused by some of the re ected signal from the test device being re ected from the

source back toward the test device and re-re ected from the test device (Figure 6-29 ). In a

transmission measurement, the source match error signal is caused by re ection from the test

device that is re-re ected from the source . Source match is most often given in terms of return

loss in dB: thus the larger the number , the smaller the error .

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 319)
Fig. 145 · Chapter 6 Application and Operation Concepts

Figure 6-29. Source Match

The error contributed by source match is dependent on the relationship between the actual

input impedance of the test device and the equivalent match of the source . It is a factor

in both transmission and re ection measurements . Source match is a particular problem in

measurements where there is a large impedance mismatch at the measurement plane . (F or

example , re ection devices such as lters with stop bands .)

Load match error results from an imperfect match at the output of the test device . It is caused

by impedance mismatches between the test device output port and port 2 of the measurement

system. As illustrated in Figure 6-30 , some of the transmitted signal is re ected from port

2 back to the test device . A portion of this wave may be re-re ected to port 2, or part may

be transmitted through the device in the reverse direction to appear at port 1. If the test

device has low insertion loss (for example a lter pass band), the signal re ected from port 2

and re-re ected from the source causes a signi cant error because the test device does not

attenuate the signal signi cantly on each re ection. Load match is usually given in terms of

return loss in dB: thus the larger the number , the smaller the error .

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 319)
Fig. 146 · Chapter 6 Application and Operation Concepts

Figure 6-30. Load Match

6-52 Application and Operation Concepts

Source Match · Load Match

Page 320 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 320 · click to enlarge
Page 320

The error contributed by load match is dependent on the relationship between the actual

output impedance of the test device and the e ective match of the return port (port 2). It is

a factor in all transmission measurements and in re ection measurements of two-port devices .

The interaction between load match and source match is less signi cant when the test device

insertion loss is greater than about 6 dB . However , source match and load match still interact

with the input and output matches of the DUT , which contributes to transmission measurement

errors . (These errors are largest for devices with highly re ective output ports .)

Leakage of energy between analyzer signal paths contributes to error in a transmission

measurement, much like directivity does in a re ection measurement. Isolation is the vector

sum of signals appearing at the analyzer samplers due to crosstalk between the reference and

test signal paths . This includes signal leakage within the test set and in both the RF and IF

sections of the receiver .

The error contributed by isolation depends on the characteristics of the test device . Isolation

is a factor in high-loss transmission measurements . However , analyzer system isolation is more

than sucient for most measurements , and correction for it may be unnecessary .

F or measuring devices with high dynamic range , accuracy enhancement can provide

improvements in isolation that are limited only by the noise oor . Generally , the isolation falls

below the noise oor , therefore , when performing an isolation calibration you should use a

noise reduction function such as averaging or reduce the IF bandwidth.

This is the vector sum of all test setup variations in which magnitude and phase change as a

function of frequency . This includes variations contributed by signal-separation devices , test

cables , adapters , and variations between the reference and test signal paths . This error is a

factor in both transmission and re ection measurements .

F or further explanation of systematic error terms and the way they are combined and

represented graphically in error models , refer to the \Characterizing Microwave Systematic

Errors" next.

Characterizing Microwave Systematic Errors

One-P ort Error Model

In a measurement of the re ection coecient (magnitude and phase) of a test device , the

measured data di ers from the actual, no matter how carefully the measurement is made .

Directivity , source match, and re ection signal path frequency response (tracking) are the

major sources of error (see Figure 6-31).

Application and Operation Concepts 6-53

Isolation (Crosstalk) · Frequency Response (Tracking)

Page 321 of the Agilent Technologies 8753D user manual — 11A
Page 321 · click to enlarge
Page 321
Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 321)
Fig. 147 · Chapter 6 Application and Operation Concepts

Figure 6-31. Sources of Error in a Re ection Measurement

T o characterize the errors , the re ection coecient is measured by rst separating the incident

signal (I) from the re ected signal (R), then taking the ratio of the two values (see Figure 6-32).

Ideally , (R) consists only of the signal re ected by the test device (S

11A

actual).

Agilent Technologies 8753D — 11A diagram (page 321)
Fig. 148

Figure 6-32. Re ection Coecient

However , all of the incident signal does not always reach the unknown (see Figure 6-33).

Some of (I) may appear at the measurement system input due to leakage through the test

set or through a signal separation device . Also , some of (I) may be re ected by imperfect

adapters between a signal separation device and the measurement plane . The vector sum of

the leakage and the miscellaneous re ections is the e ective directivity , E

. Understandably ,

DF

the measurement is distorted when the directivity signal combines vectorally with the actual

re ected signal from the unknown, S

11A

6-54 Application and Operation Concepts

, for S

Page 322 of the Agilent Technologies 8753D user manual — RF
Page 322 · click to enlarge
Page 322
Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 322)
Fig. 149 · Chapter 6 Application and Operation Concepts

Figure 6-33. E ective Directivity E

DF

Since the measurement system test port is never exactly the characteristic impedance

(50 ohms), some of the re ected signal bounces o the test port, or other impedance transitions

further down the line , and back to the unknown, adding to the original incident signal (I). This

e ect causes the magnitude and phase of the incident signal to vary as a function of S

11A

frequency . Leveling the source to produce a constant incident signal (I) reduces this error , but

since the source cannot be exactly leveled at the test device input, leveling cannot eliminate

all power variations . This re-re ection e ect and the resultant incident power variation are

caused by the source match error , E

(see Figure 6-34 ).

SF

Agilent Technologies 8753D — SF diagram (page 322)
Fig. 150

Figure 6-34. Source Match E

SF

Frequency response (tracking) error is caused by variations in magnitude and phase atness

versus frequency between the test and reference signal paths . These are due mainly to coupler

roll o , imperfectly matched samplers , and di erences in length and loss between the incident

and test signal paths . The vector sum of these variations is the re ection signal path tracking

(see Figure 6-35).

RF

Application and Operation Concepts 6-55

and · error , E

Page 323 of the Agilent Technologies 8753D user manual — DF
Page 323 · click to enlarge
Page 323
Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 323)
Fig. 151 · Chapter 6 Application and Operation Concepts

Figure 6-35. Re ection Tracking E

RF

These three errors are mathematically related to the actual data, S

, and measured data,

11A

S

, by the following equation:

11M

( S

11 A

RF

11 M

= E

E

S

D F

(1 E

11 A

S F

S

If the value of these three \E" errors and the measured test device response were known for

each frequency , the above equation could be solved for S

11A

to obtain the actual test device

response . Because each of these errors changes with frequency , their values must be known

at each test frequency . These values are found by measuring the system at the measurement

plane using three independent standards whose S

is known at all frequencies .

11A

The rst standard applied is a \perfect load," which makes S

= 0 and essentially measures

11A

directivity (see Figure 6-36). \P erfect load" implies a re ectionless termination at the

measurement plane . All incident energy is absorbed. With S

= 0 the equation can be solved

11A

, the directivity term. In practice , of course , the \perfect load" is dicult to achieve ,

DF

although very good broadband loads are available in the HP 8753D compatible calibration kits .

Agilent Technologies 8753D — DF diagram (page 323)
Fig. 152

Figure 6-36. \P erfect Load" T ermination

Since the measured value for directivity is the vector sum of the actual directivity plus

the actual re ection coecient of the \perfect load," any re ection from the termination

represents an error . System e ective directivity becomes the actual re ection coecient of the

near \perfect load" (see Figure 6-37 ). In general, any termination having a return loss value

greater than the uncorrected system directivity reduces re ection measurement uncertainty .

6-56 Application and Operation Concepts

Show 4 more labels / short text on this page
) · + · ) · for E
Page 324 of the Agilent Technologies 8753D user manual — DF
Page 324 · click to enlarge
Page 324
Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 324)
Fig. 153 · Chapter 6 Application and Operation Concepts

Figure 6-37. Measured E ective Directivity

Next, a short circuit termination whose response is known to a very high degree is used to

establish another condition (see Figure 6-38).

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 324)
Fig. 154 · Chapter 6 Application and Operation Concepts

Figure 6-38. Short Circuit T ermination

The open circuit gives the third independent condition. In order to accurately model the phase

variation with frequency due to fringing capacitance from the open connector , a specially

designed shielded open circuit is used for this step . (The open circuit capacitance is di erent

with each connector type .) Now the values for E

, directivity , E

, source match, and E

DF

SF

RF

re ection frequency response , are computed and stored (see Figure 6-39).

Application and Operation Concepts 6-57

,

Page 325 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 325 · click to enlarge
Page 325
Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 325)
Fig. 155 · Chapter 6 Application and Operation Concepts

Figure 6-39. Open Circuit T ermination

This completes the calibration procedure .

6-58 Application and Operation Concepts

Page 326 of the Agilent Technologies 8753D user manual — 11A
Page 326 · click to enlarge
Page 326

Now the unknown is measured to obtain a value for the measured response , S

11M

frequency (see Figure 6-40).

Agilent Technologies 8753D — 11M diagram (page 326)
Fig. 156

Figure 6-40. Measured S

This is the one-port error model equation solved for S

. Since the three errors and S

11A

11M

now known for each test frequency , S

can be computed as follows:

11A

D F

11 A

= S

11 M

E

S

( S

E

) + E

E

S F

11 M

D F

RF

F or re ection measurements on two-port devices , the same technique can be applied, but

the test device output port must be terminated in the system characteristic impedance . This

termination should have as low a re ection coecient as the load used to determine directivity .

The additional re ection error caused by an improper termination at the test device's output

port is not incorporated into the one-port error model.

Two-P ort Error Model

The error model for measurement of the transmission coecients (magnitude and phase) of a

two-port device is derived in a similar manner . The potential sources of error are frequency

response (tracking), source match, load match, and isolation (see Figure 6-41). These errors are

e ectively removed using the full two-port error model.

Application and Operation Concepts 6-59

Show 4 more labels / short text on this page
Device Measurement · , at each · are ·
Page 327 of the Agilent Technologies 8753D user manual — 22A
Page 327 · click to enlarge
Page 327
Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 327)
Fig. 157 · Chapter 6 Application and Operation Concepts

Figure 6-41. Major Sources of Error

The transmission coecient is measured by taking the ratio of the incident signal (I) and the

transmitted signal (T) (see Figure 6-42 ). Ideally , (I) consists only of power delivered by the

source , and (T) consists only of power emerging at the test device output.

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 327)
Fig. 158 · Chapter 6 Application and Operation Concepts

Figure 6-42. Transmission Coecient

As in the re ection model, source match can cause the incident signal to vary as a function

of test device S

. Also , since the test setup transmission return port is never exactly the

11A

characteristic impedance , some of the transmitted signal is re ected from the test set port 2,

and from other mismatches between the test device output and the receiver input, to return

to the test device . A portion of this signal may be re-re ected at port 2, thus a ecting S

21M

or part may be transmitted through the device in the reverse direction to appear at port 1,

thus a ecting S

. This error term, which causes the magnitude and phase of the transmitted

11M

signal to vary as a function of S

, is called load match, E

(see Figure 6-43).

22A

LF

6-60 Application and Operation Concepts

,

Page 328 of the Agilent Technologies 8753D user manual — XF
Page 328 · click to enlarge
Page 328
Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 328)
Fig. 159 · Chapter 6 Application and Operation Concepts

Figure 6-43. Load Match E

LF

The measured value , S

, consists of signal components that vary as a function of the

21M

as well as E

, so the input and output re ection

SF

11A

LF

22A

coecients of the test device must be measured and stored for use in the S

21A

computation. Thus , the test setup is calibrated as described above for re ection to establish

the directivity , E

, source match, E

, and re ection frequency response , E

, terms for the

DF

SF

RF

re ection measurements .

Now that a calibrated port is available for re ection measurements , the thru is connected and

, is determined by measuring the re ection coecient of the thru connection.

LF

Transmission signal path frequency response is then measured with the thru connected. The

data is corrected for source and load match e ects , then stored as transmission frequency

TF

It is very important that the exact electrical length of the thru be known.

Most calibration kits assume a zero length thru. F or some connection types

such as Type-N, this implies one male and one female port. If the test system

requires a non-zero length thru, for example , one with two male test ports , the

exact electrical delay of the thru adapter must be used to modify the built-in

calibration kit de nition of the thru.

XF

, represents the part of the incident signal that appears at the receiver without

actually passing through the test device (see Figure 6-44). Isolation is measured with the test

set in the transmission con guration and with terminations installed at the points where the

test device will be connected.

Application and Operation Concepts 6-61

Show 7 more labels / short text on this page
relationship between E · and S · and S · error-correction · load match, E · response , E · Isolation, E
Page 329 of the Agilent Technologies 8753D user manual — RF
Page 329 · click to enlarge
Page 329
Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 329)
Fig. 160 · Chapter 6 Application and Operation Concepts

Figure 6-44. Isolation E

XF

Thus there are two sets of error terms , forward and reverse , with each set consisting of six

error terms , as follows:

DF

DR

XF

XR

SF

SR

LF

LR

TF

TR

Re ection Tracking, E

RF

RR

The analyzer's test set can measure both the forward and reverse characteristics of the test

device without you having to manually remove and physically reverse the device . The full

two-port error model illustrated in Figure 6-45 depicts how the analyzer e ectively removes

both the forward and reverse error terms for transmission and re ection measurements .

6-62 Application and Operation Concepts

Show 12 more labels / short text on this page
Directivity , E · (forward) and E · (reverse) · Isolation, E · and E · Source Match, E · and E · Load Match, E · and E · Transmission Tracking, E · and E · and E
Page 330 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 330 · click to enlarge
Page 330
Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 330)
Fig. 161 · Chapter 6 Application and Operation Concepts

Figure 6-45. Full Two-P ort Error Model

Figure 6-46 shows the full two-port error model equations for all four S-parameters of a

two-port device . Note that the mathematics for this comprehensive model use all forward and

reverse error terms and measured values . Thus , to perform full error-correction for any one

parameter , all four S-parameters must be measured.

Applications of these error models are provided in the calibration procedures described in

Chapter 5, \Optimizing Measurement Results ."

Application and Operation Concepts 6-63

Page 331 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 331 · click to enlarge
Page 331
Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 331)
Fig. 162 · Chapter 6 Application and Operation Concepts

Figure 6-46. Full Two-P ort Error Model Equations

In addition to the errors removed by accuracy enhancement, other systematic errors exist

due to limitations of dynamic accuracy , test set switch repeatability , and test cable stability .

These , combined with random errors , also contribute to total system measurement uncertainty .

Therefore , after accuracy enhancement procedures are performed, residual measurement

uncertainties remain.

6-64 Application and Operation Concepts

Page 332 of the Agilent Technologies 8753D user manual — N
Page 332 · click to enlarge
Page 332

Calibration procedures are parameter-speci c , rather than channel-speci c . When a parameter

is selected, the instrument checks the available calibration data, and uses the data found

for that parameter . F or example , if a transmission response calibration is performed for

B/R, and an S

1-port calibration for A/R, the analyzer retains both calibration sets and

corrects whichever parameter is displayed. Once a calibration has been performed for a

speci c parameter or input, measurements of that parameter remain calibrated in either

channel, as long as stimulus values are coupled. In the response and response and isolation

calibrations , the parameter must be selected before calibration: other correction procedures

select parameters automatically . Changing channels during a calibration procedure invalidates

the part of the procedure already performed.

In calibration procedures that require measurement of several di erent devices , for example

a short, an open, and a load, the order in which the devices are measured is not critical.

N

NNNNNNNNNNNNN

Any standard can be re-measured, until the

DONE key is pressed. The change in trace during

measurement of a standard is normal.

Response and response and isolation calibrations require measurement of only one standard

device . If more than one device is measured, only the data for the last device is retained.

Isolation calibration can be omitted for most measurements , except where high dynamic range

is a consideration. Use the following guidelines . When the measurement requires a dynamic

range of:

90 dB: Omit isolation calibration for most measurements .

90 to 100 dBIsolation calibration is recommended with test port power greater than 0

dBm. F or this isolation calibration, averaging should be turned on with an averaging factor

at least four times the measurement averaging factor . F or example , use use an averaging

factor of 16 for the isolation calibration, and then reduce the averaging factor to four for the

measurement after calibration.

100 dB: Same as above , but alternate mode should be used. See page 5-53.

Y ou should save the calibration data, either in the internal non-volatile memory or on a disk.

If you do not save it, it will be lost if you select another calibration procedure for the same

channel, or if you change stimulus values . Instrument preset, power on, and instrument state

recall will also clear the calibration data.

Application and Operation Concepts 6-65

Show 5 more labels / short text on this page
Calibration Considerations · Measurement P arameters · Device Measurements · Omitting Isolation Calibration · Saving Calibration Data
Page 333 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 333 · click to enlarge
Page 333

During measurement calibration, the analyzer measures actual, well-de ned standards and

mathematically compares the results with ideal \models" of those standards . The di erences

are separated into error terms which are later removed during error-correction. Most of the

di erences are due to systematic errors - repeatable errors introduced by the analyzer , test set,

and cables - which are correctable .

The standard devices required for system calibration are available in compatible calibration

kits with di erent connector types . Each kit contains at least one short circuit, one open

circuit, and an impedance-matched load. In kits that require adapters for interface to

the test set ports , the adapters are phase-matched for calibration prior to measurement of

non-insertable and non-reversible devices . Other standard devices can be used by specifying

their characteristics in a user-de ned kit, as described later in this section under \Modifying

Calibration Kits ."

The accuracy improvement of the correction is limited by the quality of the standard devices ,

and by the connection techniques used. F or maximum accuracy , use a torque wrench for nal

connections .

Frequency Response of Calibration Standards

In order for the response of a reference standard to show as a dot on the smith chart display

format, it must have no phase shift with respect to frequency . Standards that exhibit such

\perfect" response are the following:

7-mm short (with no o set)

type-N male short (with no o set)

There are two reasons why other types of reference standards show phase shift after

calibration:

The reference plane of the standard is electrically o set from the mating plane of the test

port. Such devices exhibit the properties of a small length of transmission line , including a

certain amount of phase shift.

The standard is an open termination, which by de nition exhibits a certain amount of fringe

capacitance (and therefore phase shift). Open terminations which are o set from the mating

plane will exhibit a phase shift due to the o set in addition to the phase shift caused by the

fringe capacitance .

The most important point to remember is that these properties will not a ect your

measurements . The analyzer compensates for them during measurement. As a result, if these

standards are measured after a calibration, they will not appear to be \perfect" shorts or

opens . This is an indication that your analyzer is working properly and that it has successfully

performed a calibration. Figure 6-47 shows sample displays of various calibration standards

after calibration.

6-66 Application and Operation Concepts

The Calibration Standards

Page 334 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 334 · click to enlarge
Page 334

Some standards have reference planes that are electrically o set from the mating plane of the

test port. These devices will show a phase shift with respect to frequency . T able 6-3 shows

which reference devices exhibit an electrical o set phase shift. The amount of phase shift can

be calculated with the formula:

= (360 x f x l)/c where:

l = electrical length of the o set

c = speed of light (3 x 10

All open circuit terminations exhibit a phase shift over frequency due to fringe capacitance .

O set open circuits have increased phase shift because the o set acts as a small length of

transmission line . Refer to T able 6-3.

T able 6-3. Calibration Standard Types and Expected Phase Shift

T est P ort

Standard Type Expected Phase Shift

3.5-mm male O set Short

(360 f l )

3.5-mm female

2.4-mm male

2.4-mm female

3.5-mm male O set Open

(360 f l )

3.5-mm female

2.4-mm male

2.4-mm female

(360 f l )

Application and Operation Concepts 6-67

Show 28 more labels / short text on this page
Electrical O set · f = frequency · meters/second) · Fringe Capacitance · Connector Type · 7-mm Short 180 · · Type-N male · · + · Type-N female · 75 Type-N female · 7-mm Open 0 · · + · capacitance · Type N-male · · + · + · capacitance · Type N-female · O pen 0 · · + · + · capacitance · 75 Type-N female
Page 335 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 335 · click to enlarge
Page 335
Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 335)
Fig. 163 · Chapter 6 Application and Operation Concepts

Figure 6-47. Typical Responses of Calibration Standards after Calibration

6-68 Application and Operation Concepts

Page 336 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 336 · click to enlarge
Page 336

How E ective Is A ccuracy Enhancement?

The uncorrected performance of the analyzer is sucient for many measurements . However ,

the vector accuracy enhancement procedures described in Chapter 5, \Optimizing Measurement

Results ," will provide a much higher level of accuracy . Figure 6-48 through Figure 6-50

illustrate the improvements that can be made in measurement accuracy by using a more

complete calibration routine . Figure 6-48a shows a measurement in log magnitude format with

a response calibration only . Figure 6-48b shows the improvement in the same measurement

one-port calibration. Figure 6-49a shows the measurement on a Smith chart with

response calibration only , and Figure 6-49b shows the same measurement with an S

calibration.

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 336)
Fig. 164 · Chapter 6 Application and Operation Concepts

Figure 6-48. Response versus S

1-P ort Calibration on Log Magnitude F ormat

Application and Operation Concepts 6-69

using an S · one-port

Page 337 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 337 · click to enlarge
Page 337
Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 337)
Fig. 165 · Chapter 6 Application and Operation Concepts

Figure 6-49. Response versus S

1-P ort Calibration on Smith Chart

Figure 6-50 shows the response of a device in a log magnitude format, using a response

calibration in Figure 6-50a and a full two-port calibration in Figure 6-50b .

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 337)
Fig. 166 · Chapter 6 Application and Operation Concepts

Figure 6-50. Response versus Full Two-P ort Calibration

6-70 Application and Operation Concepts

Page 338 of the Agilent Technologies 8753D user manual — N
Page 338 · click to enlarge
Page 338

Correcting for Measurement Errors

key provides access to the correction menu which leads to a series of menus that

CAL

implement the error-correction concepts described in this section. A ccuracy enhancement

(error-correction) is performed as a calibration step before you measure a test device . When

key is pressed, the correction menu is displayed.

CAL

The following softkeys are located within the correction menu:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CALIBRATE MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RESUME CAL SEQUENCE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECEIVER CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL KIT [ ]

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PWRMTR CAL [OFF]

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PORT EXTENSIONS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

VELOCITY FACTOR

NNNNNNNNNNNNNNNNNNNN

SET Z

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TEST SET SW [ ]

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ALTERNATE A and B

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CHOP A and B

Ensuring a V alid Calibration

Unless interpolated error-correction is on, measurement calibrations are valid only for a

speci c stimulus state , which must be set before a calibration has begun. The stimulus state

consists of the selected frequency range , number of points , sweep time , output power , and

sweep type . Changing the frequency range , number of points , or sweep type with correction

on invalidates the calibration and turns it o . Changing the sweep time or output power

changes the status notation Cor at the left of the screen to C? , to indicate that the calibration

is in question. If correction is turned o or in question after the stimulus changes are made ,

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CORRECTION ON off recalls the original stimulus state for the current calibration.

Application and Operation Concepts 6-71

Show 5 more labels / short text on this page
The · the · CORRECTION ON off · INTERPOL on OFF · pressing
Page 339 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 339 · click to enlarge
Page 339

Y ou can activate the interpolated error-correction feature with the

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INTERPOL ON off softkey .

This feature allows you to select a subset of the frequency range or a di erent number of

points without recalibration. When interpolation is on, the system errors for the newly selected

frequencies are calculated from the system errors of the original calibration.

System performance is unspeci ed when using interpolated error-correction. The quality

of the interpolated error-correction is dependent on the amount of phase shift and the

amplitude change between measurement points . If phase shift is no greater than 180

approximately 5 measurement points , interpolated error- correction o ers a great improvement

over uncorrected measurements . The accuracy of interpolated error-correction improves as

the phase shift and amplitude change between adjacent points decrease . When you use the

analyzer in linear frequency sweep , perform the original calibration with at least 67 points per

1 GHz of frequency span for greatest accuracy with interpolated error-correction.

Interpolated error-correction is available in three sweep modes: linear frequency , power sweep ,

and CW time .

If there is a valid correction array for a linear frequency sweep , this may be

interpolated to provide correction at the CW frequency used in power sweep

or CW time modes . This correction is part of the interpolated error-correction

feature and is not speci ed.

6-72 Application and Operation Concepts

Interpolated Error-correction · · per

Page 340 of the Agilent Technologies 8753D user manual — CAL
Page 340 · click to enlarge
Page 340

There are twelve di erent error terms for a two-port measurement that can be corrected

by accuracy enhancement in the analyzer . These are directivity , source match, load match,

isolation, re ection tracking, and transmission tracking, each in both the forward and reverse

direction. The analyzer has several di erent measurement calibration routines to characterize

one or more of the systematic error terms and remove their e ects from the measured data.

The calibrate menu allows you to perform the measurement calibration routines . These

procedures range from a simple frequency response calibration to a full two-port calibration

that e ectively removes all twelve error terms .

NNNNNNNNNNNNNNNNNNNNNNNNNN

The response calibration, activated by pressing the

RESPONSE softkey within the calibrate

menu, provides a normalization of the test setup for re ection or transmission measurements .

This calibration procedure may be adequate for measurement of well matched devices . This

is the simplest error-correction to perform, and should be used when extreme measurement

accuracy is not required.

Response and Isolation Calibration

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

The response and isolation calibration, activated by pressing the

RESPONSE & ISOL'N softkey

within the calibrate menu, provides a normalization for frequency response and crosstalk

errors in transmission measurements , or frequency response and directivity errors in re ection

measurements . This procedure may be adequate for measurement of well matched high-loss

devices .

S

one-port calibration procedures , activated by pressing the

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

S11 1-PORT

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

S22 1-PORT softkey within the calibrate menu, provide directivity , source match, and

frequency response vector error-correction for re ection measurements . These procedures

provide high accuracy re ection measurements of one-port devices or properly terminated

two-port devices .

Full Two-P ort Calibration

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

The full two-port calibration, activated by pressing the

FULL 2-PORT softkey within the

calibrate menu, provides directivity , source match, load match, isolation, and frequency

response vector error-correction, in both forward and reverse directions , for transmission and

re ection measurements of two-port devices . This calibration provides the best magnitude and

phase measurement accuracy for both transmission and re ection measurements of two-port

devices , and requires an S-parameter test set.

In this type of calibration, both forward and reverse measurements must be made . Y ou have

the option of setting the ratio of the number of forward (or reverse) sweeps versus the number

NNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

of reverse (or forward) sweeps . T o access this function, press

CAL

MORE

enter the number of sweeps desired.

Application and Operation Concepts 6-73

Show 9 more labels / short text on this page
The Calibrate Menu · Response Calibration · and S · One-P ort Calibration · The S · and S · or · · TESTSET SW and
Page 341 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 341 · click to enlarge
Page 341

TRL*/LRM* Two-P ort Calibration

The TRL*/LRM* two-port calibration, activated by pressing the

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

within the calibrate menu, provides the ability to make calibrations using the TRL or LRM

method. F or more information, refer to \TRL*/LRM* Calibration," located later in this section.

6-74 Application and Operation Concepts

TRL*/LRM* 2-PORT softkey

Page 342 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 342 · click to enlarge
Page 342

If you interrupt a calibration to go to another menu, such as averaging, you can continue the

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

calibration by pressing the

RESUME CAL SEQUENCE softkey in the correction menu.

The cal kit menu provides access to a series of menus used to specify the characteristics of

calibration standards . The following softkeys are located within the cal kit menu:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SELECT CAL KIT

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SAVE USER KIT

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MODIFY [ ]

The Select Cal Kit Menu

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SELECT CAL KIT softkey within the cal kit menu provides access to the select

cal kit menu. This menu allows you to select from several default calibration kits that have

di erent connector types . These kits have prede ned standards and are valid for most

applications . It is not possible to overwrite these standard de nitions .

The numerical de nitions for most Hewlett-P ackard calibration kits can be referenced in the

calibration kit operating and service manuals , or can be viewed on the analyzer . The standard

de nitions can also be modi ed to any set of standards used.

Application and Operation Concepts 6-75

Restarting a Calibration · Cal Kit Menu · Pressing the

Page 343 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 343 · click to enlarge
Page 343

Modifying calibration kits is necessary only if unusual standards (such as in TRL*) are used

or the very highest accuracy is required. Unless a calibration kit model is provided with the

calibration devices used, a solid understanding of error-correction and the system error model

are absolutely essential to making modi cations . Y ou may use modi cations to a prede ned

calibration kit by modifying the kit and saving it as a user kit. The original prede ned

calibration kit will remain unchanged.

Before attempting to modify calibration standard de nitions , you should read application note

8510-5A to improve your understanding of modifying calibration kits . The part number of this

application note is 5956-4352. Although the application note is written for the HP 8510 family

of network analyzers , it also applies to the HP 8753D .

Several situations exist that may require a user-de ned calibration kit:

A calibration is required for a connector interface di erent from the four default calibration

kits . (ExamplesSMA, TNC, or waveguide .)

A calibration with standards (or combinations of standards) that are di erent from the

default calibration kits is required. (Example: Using three o set shorts instead of open,

short, and load to perform a 1-port calibration.)

The built-in standard models for default calibration kits can be improved or re ned.

Remember that the more closely the model describes the actual performance of the standard,

the better the calibration. (ExampleThe 7 mm load is determined to be 50.4 ohms instead

of 50.0 ohms .)

The following are de nitions of terms:

A \standard" (represented by a number 1-8) is a speci c , well-de ned, physical device used

to determine systematic errors . F or example , standard 1 is a short in the 3.5 mm calibration

kit. Standards are assigned to the instrument softkeys as part of a class .

A standard \type" is one of ve basic types that de ne the form or structure of the model to

be used with that standard (short, open, load, delay/thru, and arbitrary impedance); standard

1 is of the type short in the 3.5 mm calibration kit.

Standard \coecients" are numerical characteristics of the standards used in the model

selected. F or example , the o set delay of the short is 32 ps in the 3.5 mm calibration kit.

A standard \class" is a grouping of one or more standards that determines which of the eight

standards are used at each step of the calibration. F or example , standard number 2 and 8

usually makes up the S

A re ection class , which for type-N calibration kits are male and

female shorts .

The following steps are used to modify or de ne a user kit:

1. Select the prede ned kit to be modi ed. (This is not necessary for de ning a new calibration

kit.)

2. De ne the standards:

a. De ne which \type" of standard it is .

b. De ne the electrical characteristics (coecients) of the standard.

3. Specify the class where the standard is to be assigned.

6-76 Application and Operation Concepts

Modifying Calibration Kits · De nitions · Procedure

Page 344 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 344 · click to enlarge
Page 344
4. Store the modi ed calibration kit.

F or a step by step procedure on how to modify calibration kits , refer to \Modifying Calibration

Kit Standards" located in Chapter 5, \Optimizing Measurement Results ."

Modify Calibration Kit Menu

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MODIFY [ ] softkey in the cal kit menu provides access to the modify calibration kit

menu. This leads in turn to additional series of menus associated with modifying calibration

kits . The following is a description of the softkeys located within this menu:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DEFINE STANDARD makes the standard number the active function, and brings up the

de ne standard menus . Before selecting a standard, a standard number must be entered.

This number (1 to 8) is an arbitrary reference number used to reference standards while

specifying a class . The standard numbers for the prede ned calibration kits are as follows:

Although the numbering sequences are arbitrary , confusion can be minimized

by using consistency . However , standard 5 is always a sliding load.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SPECIFY CLASS leads to the specify class menu. After the standards are modi ed, use this

key to specify a class to group certain standards .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LABEL CLASS leads to the label class menu, to give the class a meaningful label for future

reference .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LABEL KIT leads to a menu for constructing a label for the user-modi ed cal kit. If a label

is supplied, it will appear as one of the ve softkey choices in the select cal kit menu. The

approach is similar to de ning a display title , except that the kit label is limited to ten

characters .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRL/LRM OPTION brings up the TRL Option menu.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

KIT DONE (MODIFIED) terminates the calibration kit modi cation process , after all

standards are de ned and all classes are speci ed. Be sure to save the kit with the

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SAVE USER KIT softkey , if it is to be used later .

Application and Operation Concepts 6-77

Show 9 more labels / short text on this page
The · 1 short (m) · 2 open (m) · 3 broadband load · 4 thru · 5 sliding load · 6 lowband load · 7 short (f) · 8 open (f)
Page 345 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 345 · click to enlarge
Page 345

De ne Standard Menus

Standard de nition is the process of mathematically modeling the electrical characteristics

(delay , attenuation, and impedance) of each calibration standard. These electrical

characteristics (coecients) can be mathematically derived from the physical dimensions and

material of each calibration standard, or from its actual measured response . The parameters of

the standards can be listed in T able 6-4.

T able 6-4. Standard De nitions

= Calibration Kit Label:

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 345)
Fig. 167 · Chapter 6 Application and Operation Concepts

of network analyzer is set to this value .

Open, short, load, delay/thru, or arbitrary impedance .

Load or arbitrary impedance only .

Arbitrary impedance only , device terminating impedance .

Open standard types only .

6-78 Application and Operation Concepts

Show 7 more labels / short text on this page
System Z · a · a · Ensure system Z · b · d · e
Page 346 of the Agilent Technologies 8753D user manual — N
Page 346 · click to enlarge
Page 346

Each standard must be identi ed as one of ve \types": open, short, load, delay/thru, or

arbitrary impedance .

After a standard number is entered, selection of the standard type will present one of ve

menus for entering the electrical characteristics (model coecients) corresponding to that

NNNNNNNNNNNNNN

standard type , such as

OPEN . These menus are tailored to the current type , so that only

characteristics applicable to the standard type can be modi ed.

The following is a description of the softkeys located within the de ne standard menu:

NNNNNNNNNNNNNN

OPEN de nes the standard type as an open, used for calibrating re ection measurements .

Opens are assigned a terminal impedance of in nite ohms , but delay and loss o sets may

still be added. Pressing this key also brings up a menu for de ning the open, including its

capacitance .

As a re ection standard, an open termination o ers the advantage of broadband frequency

coverage . At microwave frequencies , however , an open rarely has perfect re ection

characteristics because fringing capacitance e ects cause phase shift that varies with

frequency . This can be observed in measuring an open termination after calibration, when

an arc in the lower right circumference of the Smith chart indicates capacitive reactance .

These e ects are impossible to eliminate , but the calibration kit models include the open

termination capacitance at all frequencies for compatible calibration kits . The capacitance

model is a cubic polynomial, as a function of frequency , where the polynomial coecients

are user-de nable . The capacitance model equation is:

C = (C0) + (C1 F) + (C2 F

) + (C3 F

where F is the measurement frequency .

The terms in the equation are de ned with the specify open menu as follows:

NNNNNNNN

C0 allows you to enter the C0 term, which is the constant term of the cubic polynomial

and is scaled by 10

F arads .

NNNNNNNN

C1 allows you to enter the C1 term, expressed in F/Hz (F arads/Hz) and scaled by 10

NNNNNNNN

C2 allows you to enter the C2 term, expressed in F/Hz

and scaled by 10

N

NNNNNNN

C3 allows you to enter the C3 term, expressed in F/Hz

and scaled by 10

NNNNNNNNNNNNNNNNN

SHORT de nes the standard type as a short, for calibrating re ection measurements . Shorts

are assigned a terminal impedance of 0 ohms , but delay and loss o sets may still be added.

N

NNNNNNNNNNNNN

LOAD de nes the standard type as a load (termination). Loads are assigned a terminal

impedance equal to the system characteristic impedance Z0, but delay and loss o sets

may still be added. If the load impedance is not Z0, use the arbitrary impedance standard

de nition.

NNNNNNNNNNNNNNNNN

FIXED de nes the load as a xed (not sliding) load.

NNNNNNNNNNNNNNNNNNNNNNN

SLIDING de nes the load as a sliding load. When such a load is measured during

calibration, the analyzer will prompt for several load positions , and calculate the ideal load

value from it.

NNNNNNNNNNNNNNNNNNNN

OFFSET de nes the load as being o set.

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DELAY/THRU de nes the standard type as a transmission line of speci ed length, for

calibrating transmission measurements .

Application and Operation Concepts 6-79

Show 5 more labels / short text on this page
) · 15 · 27 · 36 · 45
Page 347 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 347 · click to enlarge
Page 347

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ARBITRARY IMPEDANCE de nes the standard type to be a load, but with an arbitrary

impedance (di erent from system Z0).

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TERMINAL IMPEDANCE allows you to specify the (arbitrary) impedance of the standard, in

ohms .

NNNNNNNNNNNNNNNNN

FIXED de nes the load as a xed (not sliding) load.

NNNNNNNNNNNNNNNNNNNNNNN

SLIDING de nes the load as a sliding load. When such a load is measured during

calibration, the analyzer will prompt for several load positions , and calculate the ideal load

value from it.

Any standard type can be further de ned with o sets in delay , loss , and standard impedance;

assigned minimum or maximum frequencies over which the standard applies; and de ned as

coax or waveguide . The

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SPECIFY OFFSET softkey provides access to the specify o set menu

(described next).

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LABEL STD softkey allows you to de ne a distinct label for each standard, so that

the analyzer can prompt the user with explicit standard labels during calibration (such as

SHORT ). The function is similar to de ning a display title , except that the label is limited to ten

characters .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

After each standard is de ned, including o sets , the

STD DONE (DEFINED) softkey will

terminate the standard de nition.

Specify O set Menu

The specify o set menu allows additional speci cations for a user-de ned standard. F eatures

speci ed in this menu are common to all ve types of standards .

O sets may be speci ed with any standard type . This means de ning a uniform length of

transmission line to exist between the standard being de ned and the actual measurement

plane . (Example: a waveguide short circuit terminator , o set by a short length of waveguide .)

F or re ection standards , the o set is assumed to be between the measurement plane and the

standard (one-way only). F or transmission standards , the o set is assumed to exist between the

two reference planes (in e ect, the o set is the thru). Three characteristics of the o set can

be de ned: its delay (length), loss , and impedance .

In addition, the frequency range over which a particular standard is valid can be de ned with

a minimum and maximum frequency . This is particularly important for a waveguide standard,

since its behavior changes rapidly beyond its cuto frequency . Note that several band-limited

standards can together be de ned as the same \class" (see specify class menu). Then, if a

measurement calibration is performed over a frequency range exceeding a single standard,

additional standards can be used for each portion of the range .

Lastly , the standard must be de ned as either coaxial or waveguide . If it is waveguide ,

dispersion e ects are calculated automatically and included in the standard model.

The following is a description of the softkeys located within the specify o set menu:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

OFFSET DELAY allows you to specify the one-way electrical delay from the measurement

(reference) plane to the standard, in seconds (s). (In a transmission standard, o set delay is

the delay from plane to plane .) Delay can be calculated from the precise physical length of

the o set, the permittivity constant of the medium, and the speed of light.

In coax, group delay is considered constant. In waveguide , however , group delay is

dispersive , that is , it changes signi cantly as a function of frequency . Hence , for a waveguide

standard, o set delay must be de ned as though it were a TEM wave (without dispersion).

6-80 Application and Operation Concepts

The

Page 348 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 348 · click to enlarge
Page 348

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

OFFSET LOSS allows you to specify energy loss , due to skin e ect, along a one-way length of

coax o set. The value of loss is entered as ohms/nanosecond (or Giga ohms/second) at 1 GHz.

(Such losses are negligible in waveguide , so enter 0 as the loss o set.)

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNN

OFFSET Z0 allows you to specify the characteristic impedance of the coax o set. (Note:

This is not the impedance of the standard itself.) F or waveguide , the o set impedance as

well as the system Z0 must always be set to 1 .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MINIMUM FREQUENCY allows you to de ne the lowest frequency at which the standard

can be used during measurement calibration. In waveguide , this must be the lower cuto

frequency of the standard, so that the analyzer can calculate dispersive e ects correctly (see

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

OFFSET DELAY above).

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MAXIMUM FREQUENCY allows you to de ne the highest frequency at which the standard can

be used during measurement calibration. In waveguide , this is normally the upper cuto

frequency of the standard.

NNNNNNNNNNNNNN

COAX de nes the standard (and the o set) as coaxial. This causes the analyzer to assume

linear phase response in any o sets .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

WAVEGUIDE de nes the standard (and the o set) as rectangular waveguide . This causes the

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

analyzer to assume a dispersive delay (see

OFFSET DELAY above).

This menu allows you to label (reference) individual standards during the menu-driven

measurement calibration sequence . The labels are user-de nable using a character set shown

on the display that includes letters , numbers , and some symbols , and they may be up to ten

characters long. The analyzer will prompt you to connect standards using these labels , so they

should be meaningful to you, and distinct for each standard.

By convention, when sexed connector standards are labeled male (m) or female (f), the

designation refers to the test port connector sex, not the connector sex of the standard.

Once a standard has been de ned, it must be assigned to a standard \class ." This is a group of

from one to seven standards that is required to calibrate for a single error term. The standards

within a single class can be assigned to the locations listed in T able 6-5 according to their

standard reference numbers .

A class often consists of a single standard, but may be composed of more than one standard if

band-limited standards are used. F or example , if there were two load standards a xed load

for low frequencies , and a sliding load for high frequencies then that class would have two

standards .

Application and Operation Concepts 6-81

Label Standard Menu · Specify Class Menu

Page 349 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNN
Page 349 · click to enlarge
Page 349

T able 6-5. Standard Class Assignments

Calibration Kit Label:

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 349)
Fig. 168 · Chapter 6 Application and Operation Concepts

The number of standard classes required depends on the type of calibration being performed,

and is identical to the number of error terms corrected. A response calibration requires only

one class , and the standards for that class may include an open, or short, or thru. A 1-port

calibration requires three classes . A full 2-port calibration requires 10 classes , not including two

for isolation.

The number of standards that can be assigned to a given class may vary from none (class not

used) to one (simplest class) to seven. When a certain class of standards is required during

calibration, the analyzer will display the labels for all the standards in that class (except when

the class consists of a single standard). This does not, however , mean that all standards in a

class must be measured during calibration. Unless band-limited standards are used, only a

single standard per class is required.

It is often simpler to keep the number of standards per class to the bare

minimum needed (often one) to avoid confusion during calibration.

Each class can be given a user-de nable label as described under label class menus .

Standards are assigned to a class simply by entering the standard's reference number

(established while de ning a standard) under a particular class . The following is a description

of the softkeys located within the specify class menu:

NNNNNNNNNNNNNN

S11A allows you to enter the standard numbers for the rst class required for an S

calibration. (F or default calibration kits , this is the open.)

6-82 Application and Operation Concepts

1-port

Page 350 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 350 · click to enlarge
Page 350

NNNNNNNNNNNNNN

S11B allows you to enter the standard numbers for the second class required for an S

1-port calibration. (F or default calibration kits , this is the short.)

NNNNNNNNNNNNNN

S11C allows you to enter the standard numbers for the third class required for an S

calibration. (F or default calibration kits , this is the load.)

NNNNNNNNNNNNNN

S22A allows you to enter the standard numbers for the rst class required for an S

calibration. (F or default calibration kits , this is the open.)

NNNNNNNNNNNNNN

S22B allows you to enter the standard numbers for the second class required for an S

1-port calibration. (F or default calibration kits , this is the short.)

NNNNNNNNNNNNNN

S22C allows you to enter the standard numbers for the third class required for an S

calibration. (F or default calibration kits , this is the load.)

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FWD TRANS. allows you to enter the standard numbers for the forward transmission thru

calibration. (F or default calibration kits , this is the thru.)

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REV TRANS. allows you to enter the standard numbers for the reverse transmission (thru)

calibration. (F or default calibration kits , this is the thru.)

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FWD MATCH allows you to enter the standard numbers for the forward match (thru)

calibration. (F or default calibration kits , this is the thru.)

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REV MATCH allows you to enter the standard numbers for the reverse match (thru)

calibration. (F or default calibration kits , this is the thru.)

NNNNNNNNNNNNNNNNNNNNNNNNNN

RESPONSE allows you to enter the standard numbers for a response calibration. This

calibration corrects for frequency response in either re ection or transmission measurements ,

depending on the parameter being measured when a calibration is performed. (F or default

kits , the standard is either the open or short for re ection measurements , or the thru for

transmission measurements .)

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RESPONSE & ISOL'N allows you to enter the standard numbers for a response & isolation

calibration. This calibration corrects for frequency response and directivity in re ection

measurements , or frequency response and isolation in transmission measurements .

N

NNNNNNNNNNNNNNNNNNNNNNNNN

TRL THRU allows you to enter the standard numbers for a TRL thru calibration.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRL REFLECT allows you to enter the standard numbers for a TRL re ect calibration.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRL LINE OR MATCH allows you to enter the standard numbers for a TRL line or match

calibration.

The label class menus are used to de ne meaningful labels for the calibration classes . These

then become softkey labels during a measurement calibration. Labels can be up to ten

characters long.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LABEL KIT softkey within the modify cal kit menu, provides access to this menu. It

is identical to the label class menu and the label standard menu described above . It allows

de nition of a label up to eight characters long.

After a new calibration kit has been de ned, be sure to specify a label for it. Choose a label

that describes the connector type of the calibration devices . This label will then appear in the

Application and Operation Concepts 6-83

Show 6 more labels / short text on this page
1-port · 1-port · 1-port · Label Class Menu · Label Kit Menu · This
Page 351 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 351 · click to enlarge
Page 351

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL KIT [ ] softkey label in the correction menu and the

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MODIFY [ ] label in the select

cal kit menu. It will be saved with calibration sets .

Once a measurement calibration has been generated with a user-de ned calibration kit, its

performance should be checked before making device measurements . T o check the accuracy

that can be obtained using the new calibration kit, a device with a well-de ned frequency

response (preferably unlike any of the standards used) should be measured. The veri cation

device must not be one of the calibration standards: measurement of one of these standards is

merely a measure of repeatability .

T o achieve more complete veri cation of a particular measurement calibration, accurately

known veri cation standards with a diverse magnitude and phase response should be

used. National standard traceable or HP standards are recommended to achieve veri able

measurement accuracy .

The published speci cations for the HP 8753D network analyzer system

include accuracy enhancement with compatible calibration kits . Measurement

calibrations made with user-de ned or modi ed calibration kits are not subject

to the HP 8753D speci cations , although a procedure similar to the system

veri cation procedure may be used.

6-84 Application and Operation Concepts

V erify performance

Page 352 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 352 · click to enlarge
Page 352

The HP 8753D RF network analyzer has the capability of making calibrations using the \TRL"

(thru-re ect-line) method. This section contains information on the following subjects:

Why Use TRL Calibration?

How TRL*/LRM* Calibration W orks

Improving Raw Source Match and Load Match F or TRL*/LRM* Calibration

The TRL Calibration Procedure

Requirements for TRL Standards

Why Use TRL Calibration?

This method is convenient in that calibration standards can be fabricated for a speci c

measurement environment, such as a transistor test xture or microstrip . Microstrip devices

in the form of chips , MMIC's , packaged transistors , or beam-lead diodes cannot be connected

directly to the coaxial ports of the analyzer . The device under test (DUT) must be physically

connected to the network analyzer by some kind of transition network or xture . Calibration

for a xtured measurement in microstrip presents additional diculties .

A calibration at the coaxial ports of the network analyzer removes the e ects of the network

analyzer and any cables or adapters before the xture; however , the e ects of the xture itself

are not accounted for . An in- xture calibration is preferable , but high-quality Short-Open-Load-

Thru (SOL T) standards are not readily available to allow a conventional Full 2-port calibration

of the system at the desired measurement plane of the device . In microstrip , a short circuit is

inductive , an open circuit radiates energy , and a high-quality purely resistive load is dicult to

produce over a broad frequency range . The Thru-Re ect-Line (TRL) 2-port calibration is an

alternative to the traditional SOL T Full 2-port calibration technique that utilizes simpler , more

convenient standards for device measurements in the microstrip environment.

F or coaxial, waveguide and other environments where high-quality impedance standards are

readily available , the traditional short, open, load, thru (SOL T) method provides the most

accurate results since all of the signi cant systematic errors are reduced. This method is

implemented in the form of the S

1-port, and full 2-port calibration selections .

In all measurement environments , the user must provide calibration standards for the desired

calibration to be performed. The advantage of TRL is that only three standards need to

be characterized as opposed to 4 in the traditional open, short, load, and thru full 2-port

calibrations . Further , the requirements for characterizing the T , R, and L standards are less

stringent and these standards are more easily fabricated.

standards used. F or example , \LRL" indicates that two lines and a re ect standard are used;

\TRM" indicates that a thru, re ection and match standards are used. All of these refer to the

same basic method.

Application and Operation Concepts 6-85

Show 5 more labels / short text on this page
TRL*/LRM* Calibration · TRL T erminology · TRL Options · 1-port, S · TRL T erminology
Page 353 of the Agilent Technologies 8753D user manual — LF
Page 353 · click to enlarge
Page 353

How TRL*/LRM* Calibration W orks

The TRL*/LRM* calibration used in the HP 8753D relies on the characteristic impedance

of simple transmission lines rather than on a set of discrete impedance standards . Since

transmission lines are relatively easy to fabricate (in a microstrip , for example), the impedance

of these lines can be determined from the physical dimensions and substrate's dielectric

constant.

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 353)
Fig. 169 · Chapter 6 Application and Operation Concepts

Figure 6-51.

HP 8753D functional block diagram for a 2-port error-corrected

F or an HP 8753D TRL* 2-port calibration, a total of 10 measurements are made to quantify

eight unknowns (not including the two isolation error terms). Assume the two transmission

leakage terms , E

, are measured using the conventional technique . The eight TRL

XF

XR

error terms are represented by the error adapters shown in Figure 6-52. Although this error

model is slightly di erent from the traditional Full 2-port 12-term model, the conventional

error terms may be derived from it. F or example , the forward re ection tracking (E

RF

represented by the product of "

. Also notice that the forward source match (E

SF

reverse load match (E

) are both represented by "

, while the reverse source match (E

LR

SR

and forward load match (E

) are both represented by "

. In order to solve for these eight

LF

unknown TRL error terms , eight linearly independent equations are required.

The rst step in the TRL* 2-port calibration process is the same as the transmission step for

a Full 2-port calibration. F or the thru step , the test ports are connected together directly

(zero length thru) or with a short length of transmission line (non- zero length thru) and the

transmission frequency response and port match are measured in both directions by measuring

all four S-parameters .

F or the re ect step , identical high re ection coecient standards (typically open or short

circuits) are connected to each test port and measured (S

).

F or the line step , a short length of transmission line (di erent in length from the thru) is

inserted between port 1 and port 2 and again the frequency response and port match are

measured in both directions by measuring all four S-parameters .

In total, ten measurements are made , resulting in ten independent equations . However , the

TRL error model has only eight error terms to solve for . The characteristic impedance of the

6-86 Application and Operation Concepts

Show 8 more labels / short text on this page
TRL* Error Model · measurement system · and E · ) is · and " · ) and · ) · and S
Page 354 of the Agilent Technologies 8753D user manual — XF
Page 354 · click to enlarge
Page 354

line standard becomes the measurement reference and, therefore , has to be assumed ideal (or

known and de ned precisely).

At this point, the forward and reverse directivity (E

DF

DR

), transmission tracking (E

TF

TR

), and re ection tracking (E

RF

RR

) terms may be derived from the TRL error

terms . This leaves the isolation (E

XF

XR

), source match (E

SF

SR

) and load match

(E

LF

LR

) terms to discuss .

Two additional measurements are required to solve for the isolation terms (E

XF

XR

).

Isolation is characterized in the same manner as the Full 2-port calibration. F orward and

reverse isolation are measured as the leakage (or crosstalk) from port 1 to port 2 with each port

terminated. The isolation part of the calibration is generally only necessary when measuring

high loss devices (greater than 70 dB).

If an isolation calibration is performed, the xture leakage must be the

same during the isolation calibration and the measurement.

Agilent Technologies 8753D — XF diagram (page 354)
Fig. 170

Figure 6-52. 8-term TRL error model and generalized coecients

Application and Operation Concepts 6-87

Show 8 more labels / short text on this page
and E · and E · and E · and E · and E · and E · Isolation · and E
Page 355 of the Agilent Technologies 8753D user manual — LF
Page 355 · click to enlarge
Page 355

Source match and load match

A TRL calibration assumes a perfectly balanced test set architecture as shown by the term

which represents both the forward source match (E

SF

) and reverse load match (E

LR

) and by the

term which represents both the reverse source match (E

SR

) and forward load match (E

LF

).

However , in any switching test set, the source and load match terms are not equal because the

transfer switch presents a di erent terminating impedance as it is changed between port 1 and

port 2.

Because the standard HP 8753D network analyzer is based on a three-sampler receiver

architecture , it is not possible to di erentiate the source match from the load match terms . The

terminating impedance of the switch is assumed to be the same in either direction. Therefore ,

the test port mismatch cannot be fully corrected. An assumption is made that:

forward source match (E

SF

) = reverse load match (E

LR

reverse source match (E

SR

) = forward load match (E

LF

F or a xture , TRL* can eliminate the e ects of the xture's loss and length, but does not

completely remove the e ects due to the mismatch of the xture .

Because the TRL technique relies on the characteristic impedance of

transmission lines , the mathematically equivalent method LRM (for

line-re ect-match) may be substituted for TRL. Since a well matched

termination is , in essence , an in nitely long transmission line , it is well suited

for low (RF) frequency calibrations . A chieving a long line standard for low

frequencies is often times physically impossible .

Improving Raw Source Match and Load Match F or TRL*/LRM* Calibration

A technique that can be used to improve the raw test port mismatch is to add high quality

xed attenuators . The e ective match of the system is improved because the xed attenuators

usually have a return loss that is better than that of the network analyzer . A dditionally , the

attenuators provide some isolation of re ected signals . The attenuators also help to minimize

the di erence between the port source match and load match, making the error terms more

equivalent.

With the attenuators in place , the e ective port match of the system is improved so that the

mismatch of the xture transition itself dominates the measurement errors after a calibration.

6-88 Application and Operation Concepts

" · ) = " · ) = "

Page 356 of the Agilent Technologies 8753D user manual — T
Page 356 · click to enlarge
Page 356
Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 356)
Fig. 171 · Chapter 6 Application and Operation Concepts

Figure 6-53. Typical Measurement Set up

If the device measurement requires bias , it will be necessary to add external bias tees between

the xed attenuators and the xture . The internal bias tees of the analyzer will not pass the

bias properly through the external xed attenuators . Be sure to calibrate with the external bias

tees in place (no bias applied during calibration) to remove their e ect from the measurement.

Because the bias tees must be placed after the attenuators , they essentially become part of the

xture . Therefore , their mismatch e ects on the measurement will not be improved by the

attenuators .

Although the xed attenuators improve the raw mismatch of the network analyzer system,

they also degrade the overall measurement dynamic range .

This e ective mismatch of the system after calibration has the biggest e ect on re ection

measurements of highly re ective devices . Likewise , for well matched devices , the e ects of

mismatch are negligible . This can be shown by the following approximation:

Re ection magnitude uncertainty = E

+ E

S

+ E

(S

+ E

S

S

D

R

S

Transmission magnitude uncertainty = E

+ E

S

+ E

S

S

+ E

S

S

T

S

where:

E

= e ective directivity

D

E

= e ective re ection tracking

R

E

= e ective source match

S

E

= e ective load match

E

= e ective crosstalk

E

= e ective transmission tracking

T

S

=S-parameters of the device under test

Application and Operation Concepts 6-89

)

Page 357 of the Agilent Technologies 8753D user manual — (MA TCH)
Page 357 · click to enlarge
Page 357

The TRL Calibration Procedure

Requirements for TRL Standards

When building a set of TRL standards for a microstrip or xture environment, the requirements

for each of these standard types must be satis ed.

No loss . Characteristic impedance (Z

) need not be known.

S

= S

S

= S

THRU

Z

of the thru must be the same as the line . (If they are not the same , the

average impedance is used.)

Attenuation of the thru need not be known.

If the thru is used to set the reference plane , the insertion phase or

electrical length must be well-known and speci ed. If a non-zero length

thru is speci ed to have zero delay , the reference plane is established in the

middle of the thru.

REFLECT Re ection coecient ( ) magnitude is optimally 1.0, but need not be

known.

Phase of must known and speci ed to within 1/4 wavelength or 90

During computation of the error model, the root choice in the solution of a

quadratic equation is based on the re ection data. An error in de nition

would show up as a 180

error in the measured phase .

must be identical on both ports .

If the re ect is used to set the reference plane , the phase response must be

well-known and speci ed.

LINE/MA TCH

Z

of the line establishes the reference impedance of the measurement

(LINE)

(i.e . S

= S

= 0). The calibration impedance is de ned to be the same as

Z

of the line . If the Z

is known but not the desired value (i.e ., not equal to

50 ), the SYSTEMS Z

selection under the TRL/LRM options menu is used.

Insertion phase of the line must not be the same as the thru (zero length

or non-zero length). The di erence between the thru and line must be

) n x 180

. Measurement uncertainty will increase

signi cantly when the insertion phase nears 0 or an integer multiple of 180

Optimal line length is 1/4 wavelength or 90

of insertion phase relative to

the thru at the middle of the desired frequency span.

Usable bandwidth for a single thru/line pair is 8:1 (frequency span:start

frequency).

Multiple thru/line pairs (Z

assumed identical) can be used to extend the

bandwidth to the extent transmission lines are available .

Attenuation of the line need not be known.

Insertion phase must be known and speci ed within

1/4 wavelength or 90

LINE/MA TCH

Z

of the match establishes the reference impedance of the measurement.

(MA TCH)

must be identical on both ports .

6-90 Application and Operation Concepts

Show 18 more labels / short text on this page
Types Requirements · THRU (Zero · length) · = 1 · · = 0 · (Non-zero · length) · · · between (20 · · and 160 · · · · ·
Page 358 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 358 · click to enlarge
Page 358

F abricating and de ning calibration standards for TRL/LRM

When calibrating a network analyzer , the actual calibration standards must have known

physical characteristics . F or the re ect standard, these characteristics include the o set in

electrical delay (seconds) and the loss (ohms/second of delay). The characteristic impedance ,

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

OFFSET Z0 , is not used in the calculations in that it is determined by the line standard. The

re ection coecient magnitude should optimally be 1.0, but need not be known since the same

re ection coecient magnitude must be applied to both ports .

The thru standard may be a zero-length or known length of transmission line . The value of

length must be converted to electrical delay , just like that done for the re ect standard. The

loss term must also be speci ed.

Application and Operation Concepts 6-91

Page 359 of the Agilent Technologies 8753D user manual — V F
Page 359 · click to enlarge
Page 359

The line standard must meet speci c frequency related criteria, in conjunction with the length

used by the thru standard. In particular , the insertion phase of the line must not be the same

as the thru. The optimal line length is 1/4 wavelength (90 degrees) relative to a zero length

thru at the center frequency of interest, and between 20 and 160 degrees of phase di erence

over the frequency range of interest. (Note: these phase values can be N 180 degrees

where N is an integer .) If two lines are used (LRL), the di erence in electrical length of

the two lines should meet these optimal conditions . Measurement uncertainty will increase

signi cantly when the insertion phase nears zero or is an integer multiple of 180 degrees , and

this condition is not recommended.

F or a transmission media that exhibits linear phase over the frequency range of interest, the

following expression can be used to determine a suitable line length of one-quarter wavelength

at the center frequency (which equals the sum of the start frequency and stop frequency

divided by 2):

E l ectr ical l eng th (cm) = (LI N E 0 l eng th T H R U )

(15000 V F )

E l ectr ical l eng th (cm) =

f 1(M H z ) + f 2(M H z )

let:

f1 = 1000 MHz

f2 = 2000 MHz

VF = V elocity F actor = 1 (for this example)

Thus , the length to initially check is 5 cm.

Next, use the following to verify the insertion phase at f1 and f2:

(360 f l )

P hase (deg r ees ) =

where:

l = length of line

v = velocity = speed of light velocity factor

which can be reduced to the following using frequencies in MHz and length in centimeters:

0. 012 f ( M H z ) l ( cm )

P hase (deg r ees ) appr ox =

V F

So for an air line (velocity factor approximately 1) at 1000 MHz, the insertion phase is

60 degrees for a 5 cm line; it is 120 degrees at 2000 MHz. This line would be a suitable line

standard.

F or microstrip and other fabricated standards , the velocity factor is signi cant. In those cases ,

the phase calculation must be divided by that factor . F or example , if the dielectric constant for

a substrate is 10, and the corresponding \e ective" dielectric constant for microstrip is 6.5,

then the \e ective" velocity factor equals 0.39 (1 square root of 6.5).

Using the rst equation with a velocity factor of 0.39, the initial length to test would be 1

.95 cm. This length provides an insertion phase at 1000 MHz of 60 degrees; at 2000 MHz,

120 degrees (the insertion phase should be the same as the air line because the velocity factor

was accounted for when using the rst equation).

6-92 Application and Operation Concepts

f = frequency

Page 360 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 360 · click to enlarge
Page 360

Another reason for showing this example is to point out the potential problem in calibrating at

low frequencies using TRL. F or example , one-quarter wavelength is

7500 V F

Leng th (cm) =

where:

fc = center frequency

Thus , at 50 MHz,

Leng th (cm) =

= 150 cm or 15 m

50 (M H z )

Such a line standard would not only be dicult to fabricate , but its long term stability and

usability would be questionable as well.

Thus at lower frequencies and/or very broad band measurements , fabrication of a \match" or

termination may be deemed more practical. Since a termination is , in essence , an in nitely

long transmission line , it ts the TRL model mathematically , and is sometimes referred to as a

\TRM" calibration.

The TRM calibration technique is related to TRL with the di erence being that it bases the

characteristic impedance of the measurement on a matched Z

termination instead of a

O

transmission line for the third measurement standard. Like the TRL thru standard, the TRM

THRU standard can either be of zero length or non-zero length. The same rules for thru and

re ect standards used for TRL apply for TRM.

TRM has no inherent frequency coverage limitations which makes it more convenient in some

measurement situations . A dditionally , because TRL requires a di erent physical length for the

thru and the line standards , its use becomes impractical for xtures with contacts that are at a

xed physical distance from each other .

F or information on how to modify calibration constants for TRL*/LRM*, and how to perform a

TRL or TRM calibration, refer to Chapter 5, \Optimizing Measurement Results ."

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRL/LRM OPTION softkey provides access to the TRL/LRM options menu. There are two

selections under this menu:

NNNNNNNNNNNNNNNNNNNNNNN

CAL ZO(calibration Z

N

SET REF(set reference)

NNNNNNNNNNNNNNNNNNNNNNNNN

The characteristic impedance used during the calibration can be referenced to either the line

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

(or match) standard (

CAL ZOLINE ZO ) or to the system (
CAL ZOSYSTEM ZO ). The analyzer

defaults to a calibration impedance that is equal to the line (or match) standard.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL ZOLINE ZO is selected, the impedance of the line (or match) standard is

assumed to match the system impedance exactly (the line standard is re ectionless). After

a calibration, all measurements are referenced to the impedance of the line standard. F or

example , when the line standard is remeasured, the response will appear at the center of the

Smith chart. When

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL ZOLINE ZO is selected, the values entered for

NNNNNNNNNNNNNNNNNNNN

SET ZO (under CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

OFFSET ZO (within the de ne standard menu) are ignored.

Application and Operation Concepts 6-93

Show 6 more labels / short text on this page
f c · TRL Options · The · ) · When the · menu) and
Page 361 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 361 · click to enlarge
Page 361

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL ZOSYSTEM ZO is selected when the desired measurement impedance di ers from the

impedance of the line standard. This requires a knowledge of the exact value of the Z

NNNNNNNNNNNNNNNNNNNN

the line . The system reference impedance is set using

SET ZO under the calibration menu.

The actual impedance of the line is set by entering the real part of the line impedance as the

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

OFFSET ZO within the de ne standard menu. F or example , if the line was known to have

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

a characteristic impedance of 51 (

OFFSET ZO = 51 ), it could still be used to calibrate

for a 50 measurement (

NNNNNNNNNNNNNNNNNNNN

SET ZO = 50 ). After a calibration, all measurements would be

referenced to 50 , instead of 51 . When the line standard is remeasured, the center of the

NNNNNNNNNNNNNNNNNNNN

Smith chart is at the current value of

SET ZO (in this case , 50 ). Since only one value of

can be selected for the line standard, the value of Z

should be a constant value over

the frequency range of interest in order to be meaningful.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

The location of the reference plane is determined by the selection of

SET REFTHRU and

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SET REFREFLECT . By default, the reference plane is set with the thru standard which

must have a known insertion phase or electrical length. If a non-zero length thru is speci ed

to have zero delay , the reference plane will be established in the middle of the thru. The

re ect standard may be used to set the reference plane instead of the thru provided the phase

response (o set delay , reactance values and standard type) of the re ect standard is known

and is speci ed in the calibration kit de nition.

Dispersion occurs when a transmission medium exhibits a variable propagation

or phase velocity as a function of frequency . The result of dispersion is a

non-linear phase shift versus frequency , which leads to a group delay which

is not constant. F ortunately , the TRL calibration technique accounts for

dispersive e ects of the test xture up to the calibration plane , provided that:

1. The thru (zero or non-zero length) is de ned as having zero electrical length

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

and is used to set the reference plane (

SET REF: THRU ).

2. The transmission lines used as calibration standards have identical dispersion

characteristics (i.e ., identical height, width and relative dielectric constant).

When a non-zero length thru is used to set the reference plane , it should be

de ned as having zero length in the TRL standards de nition, even though it

has physical length. The actual electrical length of the thru standard must then

be subtracted from the actual electrical length of each line standard in the TRL

calibration kit de nition. The device must then be mounted between two short

lengths of transmission line so that each length is exactly one-half of the length

of the non-zero length thru standard. In this con guration, the measurement

will be properly calibrated up to the point of the device .

6-94 Application and Operation Concepts

of · o set Z · Dispersion E ects

Page 362 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 362 · click to enlarge
Page 362

P ower Meter Calibration

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PWRMTR CAL [ ] softkey within the correction menu, leads to a series of menus

associated with power meter calibration.

An HP-IB-compatible power meter can monitor and correct RF source power to achieve leveled

power at the test port. During a power meter calibration, the power meter samples the power

at each measurement point across the frequency band of interest. The analyzer then constructs

a correction data table to correct the power output of the internal source . The correction table

may be saved in an instrument state register with the

NNNNNNNNNNNNNN

SAVE key .

The correction table may be updated on each sweep (in a leveling application) or during

an initial single sweep . In the sample-and-sweep mode the power meter is not needed for

subsequent sweeps . The correction table may be read or modi ed through HP-IB .

when you are testing a system with signi cant frequency response errors ( F or example , a

coupler with signi cant roll-o , or a long cable with a signi cant amount of loss .)

when you are measuring devices that are very sensitive to actual input power for proper

when you require a reference for receiver power calibration

Calibrated P ower Level

By setting the analyzer calibrated power to the desired value at the power meter , this power

level will be maintained at that port during the entire sweep . First set the source power so that

the power at the test device is approximately correct. This reduces residual power errors when

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

only one reading is taken. Refer to

NUMBER OF READINGS softkey description in

Chapter 9, \Key De nitions ." When power meter calibration is on, the annotation PC is

displayed. This indicates that the source power is being updated during the sweep . Calibrated

power level becomes the active entry if any of the following softkeys are pressed:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PWRMTR CAL [OFF]

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EACH SWEEP

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ONE SWEEP

NNNNNNNNNNNNNNNNN

POWER (if power meter cal is on)

Regardless of the measurement application, the analyzer's source can only supply corrected

power within the selected power range . If power outside this range is requested, the

annotation will change to PC? .

P ower meter calibration may be used in linear , log, list, CW , and power sweep modes . In power

sweep , the power at each point is the true power at the power meter , not the power at the

analyzer's source output.

Application and Operation Concepts 6-95

Show 4 more labels / short text on this page
The · Primary Applications · operation · Compatible Sweep Types
Page 363 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 363 · click to enlarge
Page 363

Loss of P ower Meter Calibration Data

The power meter calibration data will be lost by committing any of the following actions:

Turning power o . Turning o the instrument erases the power meter calibration table .

Changing sweep type . If the sweep type is changed (linear , log, list, CW , power) while

power meter calibration is on, the calibration data will be lost. However , calibration data is

retained if you change the sweep type while power meter calibration is o .

Changing frequency . P ower meter calibration data will also be lost if the frequency is

changed in log or list mode , but it is retained in linear sweep mode .

PRESET

. Presetting the instrument will erase power meter calibration data. If the

instrument state has been saved in a register using the

SA VE/RECALL

key , you may recall the

instrument state and the data will be restored. Saving the instrument state will not protect

the data if the instrument is turned o .

Interpolation in P ower Meter Calibration

If the frequency is changed in linear sweep , or the start/stop power is changed in power sweep ,

then the calibration data is interpolated for the new range .

If calibration power is changed in any of the sweep types , the values in the power setting array

are increased or decreased to re ect the new power level. Some accuracy is lost when this

occurs .

P ower Meter Calibration Modes of Operation

Continuous Sample Mode (Each Sweep)

Y ou can set the analyzer to update the correction table at each point for sweep (as in a

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

leveling application), using the

EACH SWEEP softkey . In this mode , the analyzer checks the

power level at every frequency point each time it sweeps . Y ou can also have more than one

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

sample/correction iteration at each frequency point. (See the

NUMBER OF READINGS softkey

description in Chapter 9.)

While using the continuous sample mode , the power meter must remain connected as shown

in Figure 6-54. A power splitter or directional coupler samples the actual power going to the

test device and is measured by the power meter . The power meter measurement provides the

information necessary to update the correction table via HP-IB .

Continuous correction slows the sweep speed considerably , especially when low power levels

are being measured by the power meter . It may take up to 10 seconds per point if the power

level is less than 20 dBm. F or faster operation, you can use the sample-and-sweep mode . If

you use a directional coupler , you must enter the attenuation of the coupled arm with respect

to the through arm using the

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

POWER LOSS softkey .

6-96 Application and Operation Concepts

Pressing

Page 364 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 364 · click to enlarge
Page 364

Figure 6-54. T est Setup for Continuous Sample Mode

Sample-and-Sweep Mode (One Sweep)

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Y ou can use the

ONE SWEEP softkey to activate the sample-and-sweep mode . This will correct

the analyzer output power and update the power meter calibration data table during the initial

measurement sweep . In this mode of operation, the analyzer does not require the power meter

for subsequent sweeps . Y ou may use a power splitter or directional coupler , or simply connect

the power sensor directly to the analyzer to measure the power for the initial sweep prior to

connecting and measuring the test device (see Figure 6-55).

The speed of the calibration will be slow while power meter readings are taken (see T able 6-6 ).

However , once the sample sweep is nished, subsequent sweeps are power-corrected using

the data table , and sweep speed increases signi cantly . Once the initial sweep is taken,

sample-and-sweep correction is much faster than continuous sample correction.

If the calibrated power level is changed after the initial measurement sweep is done , the entire

correction table is increased or decreased by that amount and the annotation PCo (indicating

power calibration o set) appears on the display . The resulting power will no longer be as

accurate as the original calibration.

Figure 6-55. T est Setup for Sample-and-Sweep Mode

Application and Operation Concepts 6-97

Page 365 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 365 · click to enlarge
Page 365

P ower Loss Correction List

If a directional coupler or power splitter is used to sample the RF power output of the analyzer ,

the RF signal going to the power meter may be di erent than that going to the test device . A

directional coupler will attenuate the RF signal by its speci ed coupling factor . The di erence

in attenuation between the through arm and the coupled arm (coupling factor) must be entered

using the loss/sensor list menu. Non-linearities in either the directional coupler or power

splitter can be corrected in the same way .

P ower loss information is entered in much the same way as limit line parameters . Up to 12

segments may be entered, each with a di erent attenuation value . The entered data will not

be lost if the instrument's power is cycled.

P ower Sensor Calibration F actor List

Two power sensor calibration data lists can be created in the analyzer . No single power sensor

covers the entire frequency range of the analyzer , therefore the calibration data for two

di erent power sensors must be available . The entered data will not be lost if the instrument's

power is cycled.

Speed and A ccuracy

The speed and accuracy of a power meter calibration vary depending on the test setup and the

measurement parameters . F or example , number of points , number of readings , if the power is

less than 20 dBm, continuous versus sample and sweep mode . A ccuracy is improved if you

set the source power such that it is approximately correct at the measurement port. P ower

meter calibration should then be turned on. With number of readings = 2, very accurate

measurements are achieved.

T able 6-6 shows typical sweep speed and power accuracy . The times given apply only to the

test setup for continuous correction or for the rst sweep of sample-and-sweep correction.

The typical values given in T able 6-6 were derived under the following conditions:

T est Equipment Used

HP 8753D network analyzer

HP 436A power meter

HP 8485A power sensor

The time required to perform a power meter calibration depends on the source power and

number of points tested. The parameters used to derive the typical values in T able 6-6 are as

follows:

number of points51, 100 kHz to 3 GHz
test port powerequal to calibration power

Sweep time is linearly proportional to the number of points measured. F or example , a sweep

taking 49 seconds at 51 points will take approximately 98 seconds if 101 points are measured.

6-98 Application and Operation Concepts

Stimulus P arameters

Page 366 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 366 · click to enlarge
Page 366

T able 6-6.

Characteristic P ower Meter Calibration Sweep Speed and A ccuracy

Number of Readings Sweep Time

at T est P ort (dBm)

1 33 0.7

+5 2 64 0.2

3 95 0.1

1 48 0.7

15 2 92 0.2

3 123 0.1

1 194 0.7

30 2 360 0.2

3 447 0.1

1 Sweep speed applies to every sweep in continuous correction mode , and

to the rst sweep in sample-and-sweep mode . Subsequent sweeps in

sample-and-sweep mode will be much faster .

2 The accuracy values were derived by combining the accuracy of the

power meter and linearity of the analyzer's internal source , as well as

the mismatch uncertainty associated with the power sensor .

The accuracy values in T able 6-6 were derived by combining the accuracy of the power meter

and linearity of the analyzer's internal source , as well as the mismatch uncertainty associated

with the test set and the power sensor .

P ower meter calibration measures the source power output (at the measurement port) at a

single stimulus point, and compares it to the calibrated power you selected. If the two values

are di erent, power meter calibration changes the source output power by the di erence . This

process is repeated at every stimulus point. The accuracy of the result depends on the amount

of correction required. If the selected number of readings = 1, the nal measurement accuracy

is signi cantly a ected by a large power change . However , if the selected number of readings

is >1, the power change on the second or third reading is much smaller: thus accuracy is much

better .

Set source power approximately correct at the measurement port, then activate power meter

calibration. This method can signi cantly increase the accuracy of the measurement when the

selected number of readings = 1. Smaller accuracy improvements occur with a higher number

of readings . Remember that mismatch errors a ect accuracy as well.

P ower meter correction applies to one port only; the other port is not

corrected.

Application and Operation Concepts 6-99

Show 4 more labels / short text on this page
P ower Desired · Characteristic · (seconds) · A ccuracy (dB)
Page 367 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 367 · click to enlarge
Page 367

Alternate and Chop Sweep Modes

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Y ou can select the

ALTERNATE A and B or

CHOP A and B softkey within the Correction More

menu to activate either one or the other sweep modes . F or information about sweep types ,

refer to \Sweep Type Menu," located earlier in this chapter .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ALTERNATE A and B measures only one input per frequency sweep , in order to reduce

unwanted signals , such as crosstalk from sampler A to B when measuring B/R. Thus , this mode

optimizes the dynamic range for all four S-parameter measurements .

The disadvantages of this mode are associated with simultaneous transmission/re ection

measurements or full two-port calibrations: this mode takes twice as long as the chop mode to

make these measurements .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CHOP A and B is the default measurement mode . This mode measures both inputs A and B

during each sweep . Thus , if each channel is measuring a di erent parameter and both channels

are displayed, the chop mode o ers the fastest measurement time . This is the preferred

measurement mode for full two-port calibrations because both inputs remain active .

The disadvantage of this mode is that in measurements of high rejection devices greater than

85 dB , such as lters with a low-loss passband, maximum dynamic range may not be achieved.

Figure 6-56 shows the alternate sweep mode (bold trace) overlaying the chop sweep mode in

a band-pass lter measurement. Note the di erence in the crosstalk levels between the two

modes .

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN diagram (page 367)
Fig. 172

Figure 6-56. Alternate and Chop Sweeps Overlaid

6-100 Application and Operation Concepts

Alternate · Chop

Page 368 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 368 · click to enlarge
Page 368

Calibrating for Non-Insertable Devices

A test device having the same sex connector on both the input and output cannot be

connected directly into a transmission test con guration. Therefore , the device is considered

to be noninsertable, and one of the following calibration methods must be performed.

F or information on performing measurement calibrations , refer to Chapter 5, \Optimizing

Measurement Results ."

The adapter removal technique provides a means to accurately measure the noninsertable

device . F or each port, a separate 2-port error correction needs to be performed to create

a calibration set. The adapter removal algorithm uses the resultant data from the two

calibration sets and the nominal electrical length of the adapter to compute the adapters actual

S-parameters . This data is then used to generate a separate third cal set in which the forward

and reverse match and tracking terms are as if P ort 1 and P ort 2 could be connected. This is

possible because the actual S-parameters of the adapter are measured with great accuracy ,

thus allowing the e ects of the adapter to be completely removed when the third cal set is

generated. See Chapter 5.

With this method, you use two precision matched adapters which are \equal." T o be equal, the

adapters must have the same match, Z

, insertion loss , and electrical delay .

Modify the Cal Kit Thru De nition

With this method it is only necessary to use one adapter . The calibration kit thru de nition is

modi ed to compensate for the adapter and then saved as a user kit. However , the electrical

delay of the adapter must rst be found.

Application and Operation Concepts 6-101

A dapter Removal · Matched A dapters

Page 369 of the Agilent Technologies 8753D user manual — SA VE/RECALL
Page 369 · click to enlarge
Page 369

Using the Instrument State Functions

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 369)
Fig. 173 · Chapter 6 Application and Operation Concepts

Figure 6-57. Instrument State Function Block

The instrument state function block keys provide control of channel-independent system

functions . The following keys are described in this chapter:

SYSTEM

: Limit lines and limit testing, time domain operation, and instrument modes .

LOCAL

: HP-IB controller modes , instrument addresses , and the use of the parallel port.

SEQ

: T est sequencing.

Information on the remaining instrument state keys can be found in the following chapters:

PRESET

Chapter 12, \Preset State and Memory Allocation"

COPY

Chapter 4, \Printing, Plotting, and Saving Measurement Results"

SA VE/RECALL

Chapter 4, \Printing, Plotting, and Saving Measurement Results"

6-102 Application and Operation Concepts

Show 6 more labels / short text on this page
· · · · ·
Page 370 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 370 · click to enlarge
Page 370

This section contains information on the following topics:

using the parallel port

LOCAL

This key is allows you to return the analyzer to local (front panel) operation from remote

(computer controlled) operation. This key will also abort a test sequence or hardcopy

print/plot. In this local mode , with a controller still connected on HP-IB , you can operate the

analyzer manually (locally) from the front panel. This is the only front panel key that is not

disabled when the analyzer is remotely controlled over HP-IB by a computer . The exception to

this is when local lockout is in e ect: this is a remote command that disables the

LOCAL

making it dicult to interfere with the analyzer while it is under computer control.

In addition, the

key provides access to the HP-IB menu, where you can set the

LOCAL

controller mode , and to the address menu, where you can enter the HP-IB addresses of

peripheral devices and select plotter/printer ports . Y ou can also set the mode of the parallel

port here .

The HP-IB menu consists of the following softkeys:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SYSTEM CONTROLLER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TALKER/LISTENER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

USE PASS CONTROL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SET ADDRESS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PARALLEL [ ]

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

HP-IB DIAG on OFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DISK UNIT NUMBER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

VOLUME NUMBER

The analyzer is factory-equipped with a remote programming interface using the

Hewlett-P ackard Interface Bus (HP-IB). This enables communication between the analyzer and

a controlling computer as well as other peripheral devices . This menu indicates the present

HP-IB controller mode of the analyzer . Three HP-IB modes are possible: system controller ,

talker/listener , and pass control.

Application and Operation Concepts 6-103

Show 7 more labels / short text on this page
HP-IB Menu · local key · HP-IB controller modes · instrument addresses · · K ey · key ,
Page 371 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 371 · click to enlarge
Page 371

HP-IB ST A TUS Indicators

When the analyzer is connected to other instruments over HP-IB , the HP-IB ST A TUS indicators

in the instrument state function block light up to display the current status of the analyzer .

R = remote operation

L = listen mode

T = talk mode

S = service request (SRQ) asserted by the analyzer

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SYSTEM CONTROLLER softkey activates the system controller mode . When in this mode ,

the analyzer can use HP-IB to control compatible peripherals , without the use of an external

computer . It can output measurement results directly to a compatible printer or plotter , store

instrument states using a compatible disk drive , or control a power meter for performing

service routines . The power meter calibration function requires system controller or pass

control mode .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TALKER/LISTENER softkey activates the talker/listener mode , which is the mode of

operation most often used. In this mode , a computer controller communicates with the

analyzer and other compatible peripherals over the bus . The computer sends commands or

instructions to and receives data from the analyzer . All of the capabilities available from the

analyzer front panel can be used in this remote operation mode , except for control of the

power line switch and some internal tests .

P ass Control Mode

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

USE PASS CONTROL softkey activates the third mode of HP-IB operation: the pass control

mode . In an automated system with a computer controller , the controller can pass control of

the bus to the analyzer on request from the analyzer . The analyzer is then the controller of the

peripherals , and can direct them to plot, print, or store without going through the computer .

When the peripheral operation is complete , control is passed back to the computer . Only one

controller can be active at a time . The computer remains the system controller , and can regain

control at any time .

Preset does not a ect the selected controller mode , but cycling the power returns the analyzer

to talker/listener mode .

Information on compatible peripherals is provided in Chapter 11, \Compatible P eripherals ."

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

This menu can be accessed by pressing the

SET ADDRESS softkey within the HP-IB menu.

In communications through the Hewlett-P ackard Interface Bus (HP-IB), each instrument on the

bus is identi ed by an HP-IB address . This decimal-based address code must be di erent for

each instrument on the bus .

This menu lets you set the HP-IB address of the analyzer , and enter the addresses of peripheral

devices so that the analyzer can communicate with them.

6-104 Application and Operation Concepts

Show 6 more labels / short text on this page
System Controller Mode · The · T alker/Listener Mode · The · The · A ddress Menu
Page 372 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNN
Page 372 · click to enlarge
Page 372

Most of the HP-IB addresses are set at the factory and need not be modi ed for normal system

operation. The standard factory-set addresses for instruments that may be part of the system

are as follows:

Instrument HP-IB A ddress

External Disk Drive 00

P ower Meter 13

The address displayed in this menu for each peripheral device must match the address set on

the device itself. The analyzer does not have an HP-IB switch: its address is set only from the

front panel.

These addresses are stored in non-volatile memory and are not a ected by preset or by cycling

the power .

Using the P arallel P ort

The instrument's parallel port can be used in two di erent modes . By pressing

LOCAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNN

PARALLEL [ ] softkey , you can select either the

[COPY] mode or the

NNNNNNNNNNNNNNNNNNNN

[GPIO] mode .

The copy mode allows the parallel port to be connected to a printer or plotter for the

outputting of test results . T o use the parallel port for printing or plotting, you must do the

following:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOCAL

SET ADDRESSES .

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRINTER PORT .

NNNNNNNNNNNNNNNNNNNNNNNNNN

PARALLEL so that copy is underlined.

The GPIO mode turns the parallel port into a \general purpose input/output" port.

In this mode the port can be connected to test xtures , power supplies , and other peripheral

equipment that might be used to interact with the analyzer during measurements . This mode is

exclusively used in test sequencing.

Application and Operation Concepts 6-105

Show 14 more labels / short text on this page
(decimal) · Analyzer 16 · Plotter 05 · Printer 01 · Controller 21 · and · then toggling the · The Copy Mode · Press · · Select either · PLOTTER PORT or · Select · The GPIO Mode
Page 373 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 373 · click to enlarge
Page 373

key provides access to the system menu. This menu leads to additional menus

SYSTEM

which control various aspects of the analyzer system. The following softkeys are located

within the system menu:

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNN

SET CLOCK allows you to produce time stamps on plots and print-outs .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LIMIT MENU provides access to the limits menu.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRANSFORM MENU (Option 010 Only) provides access to the transform menu.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

HARMONIC MEAS (Option 002 Only) provides access to the harmonic mode menu.

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INSTRUMENT MODE provides access to the instrument mode menu.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SERVICE MENU provides access to the service menu (see the HP 8753D Network Analyzer

Service Guide).

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

This menu can be accessed by pressing

LIMIT MENU softkey within the system menu.

Y ou can have limit lines drawn on the display to represent upper and lower limits or device

speci cations with which to compare the test device . Limits are de ned in segments , where

each segment is a portion of the stimulus span. Each limit segment has an upper and a lower

starting limit value . Three types of segments are available: at line , sloping line , and single

point.

Limits can be de ned independently for the two channels , up to 22 segments for each channel.

These can be in any combination of the three limit types .

Limit testing compares the measured data with the de ned limits , and provides pass or fail

information for each measured data point. An out-of-limit test condition is indicated in ve

ways: with a FAIL message on the screen, with a beep , by changing the color of the failing

portions of a trace , with an asterisk in tabular listings of data, and with a bit in the HP-IB

event status register B . (The analyzer also has a BNC rear panel output that includes this status ,

but is only valid for a single channel measurement.)

The limit test output has three selectable modes . F or more information, refer to

Chapter 2, \Making Measurements ."

Limit lines and limit testing can be used simultaneously or independently . If limit lines are

on and limit testing is o , the limit lines are shown on the display for visual comparison and

adjustment of the measurement trace . However , no pass/fail information is provided. If limit

testing is on and limit lines are o , the speci ed limits are still valid and the pass/fail status is

indicated even though the limit lines are not shown on the display .

Limits are entered in tabular form. Limit lines and limit testing can be either on or o while

limits are de ned. As new limits are entered, the tabular columns on the display are updated,

and the limit lines (if on) are modi ed to the new de nitions . The complete limit set can be

o set in either stimulus or amplitude value .

Limits are checked only at the actual measured data points . It is possible for a device to be out

of speci cation without a limit test failure indication if the point density is insucient. Be sure

to specify a high enough number of measurement points in the stimulus menu.

6-106 Application and Operation Concepts

The System Menu · The · The Limits Menu

Page 374 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNN
Page 374 · click to enlarge
Page 374

Limit lines are displayed only on Cartesian formats . In polar and Smith chart formats , limit

testing of one value is available: the value tested depends on the marker mode and is the

magnitude or the rst value in a complex pair . The message NO LIMIT LINES DISPLAYED is

shown on the display in polar and Smith chart formats .

The list values feature in the copy menu provides tabular listings to the display or a printer

for every measured stimulus value . These include limit line or limit test information if these

functions are activated. If limit testing is on, an asterisk is listed next to any measured value

that is out of limits . If limit lines are on, and other listed data allows sucient space , the upper

limit and lower limit are listed, together with the margin by which the device data passes or

fails the nearest limit.

If limit lines are on, they are plotted with the data on a plot. If limit testing is on, the PASS or

FAIL message is plotted, and the failing portions of the trace that are a di erent color on the

display are also a di erent color on the plot. If limits are speci ed, they are saved in memory

with an instrument state .

This menu allows you to specify limits for limit lines or limit testing, and presents a table of

limit values on the display . Limits are de ned in segments . Each segment is a portion of the

stimulus span. Up to 22 limit segments can be speci ed for each channel. The limit segments

do not have to be entered in any particular order: the analyzer automatically sorts them and

lists them on the display in increasing order of start stimulus value .

F or each segment, the table lists the segment number , the starting stimulus value , upper limit,

lower limit, and limit type . The ending stimulus value is the start value of the next segment,

or a segment can be terminated with a single point segment. Y ou can enter limit values as

upper and lower limits or delta limits and middle value . As new limit segments are de ned, the

tabular listing is updated. If limit lines are switched on, they are shown on the display .

If no limits have been de ned, the table of limit values shows the notation EMPTY . Limit

NNNNNNNNNNN

NNNNNNNNNNNNNN

segments are added to the table using the

ADD softkey or edited with the

EDIT softkey , as

previously described. The last segment on the list is followed by the notation END .

This menu sets the values of the individual limit segments . The segment to be modi ed, or a

default segment, is selected in the edit limits menu. The stimulus value can be set with the

controls in the entry block or with a marker (the marker is activated automatically when this

menu is presented). The limit values can be de ned as upper and lower limits , or delta limits

and middle value . Both an upper limit and a lower limit (or delta limits) must be de ned:

if only one limit is required for a particular measurement, force the other out of range (for

example +500 dB or 500 dB).

As new values are entered, the tabular listing of limit values is updated.

Segments do not have to be listed in any particular order: the analyzer sorts them

automatically in increasing order of start stimulus value when the

NNNNNNNNNNNNNN

DONE key in the edit limits

menu is pressed. However , the easiest way to enter a set of limits is to start with the lowest

stimulus value and de ne the segments from left to right of the display , with limit lines turned

on as a visual check.

Phase limit values can be speci ed between +500

. Limit values above +180

are mapped into the range of 180

to correspond with the range of

phase data values .

Application and Operation Concepts 6-107

Show 12 more labels / short text on this page
Edit Limits Menu · Edit Segment Menu · · and 500 · · · and · below 180 · · · to +180 ·
Page 375 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 375 · click to enlarge
Page 375

O set Limits Menu

This menu allows the complete limit set to be o set in either stimulus value or amplitude

value . This is useful for changing the limits to correspond with a change in the test setup , or

for device speci cations that di er in stimulus or amplitude . It can also be used to move the

limit lines away from the data trace temporarily for visual examination of trace detail.

6-108 Application and Operation Concepts

Page 376 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 376 · click to enlarge
Page 376

Knowing the Instrument Modes

There are ve major instrument modes of the analyzer:

frequency o set operation

harmonic mode operation (Option 002)

The instrument mode menu can be accessed by pressing

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SYSTEM

INSTRUMENT MODE . This

menu contains the following softkeys:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NETWORK ANALYZER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EXT SOURCE AUTO

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EXT SOURCE MANUAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TUNED RECEIVER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FREQ OFFS MENU

This is the standard mode of operation for the analyzer , and is active after you press

PRESET

or switch on the A C power .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SYSTEM

INSTRUMENT MODE

NETWORK ANALYZER returns the analyzer to the \normal"

network analyzer operating mode .

This mode uses the analyzer's internal source .

This mode allows the analyzer to phase lock to an external CW signal. External source mode is

best used for unknown signals , or for signals that drift. If a synthesized external source is used,

the tuned receiver mode is recommended because it is faster .

External source mode is useful in several applications:

when your test device is a mixer or other frequency translation device

in automated test applications where a source is already connected to the system, and you do

not want to switch between the system source and the analyzer's internal source .

Application and Operation Concepts 6-109

Show 10 more labels / short text on this page
network analyzer mode · external source mode · tuned receiver mode · · Network Analyzer Mode · · Pressing · · External Source Mode · Primary Applications
Page 377 of the Agilent Technologies 8753D user manual — SYSTEM
Page 377 · click to enlarge
Page 377

Typical T est Setup

Figure 6-58 shows a typical test setup using the external source mode . The same test setup is

applicable for either manual or automatic external source mode operation.

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 377)
Fig. 174 · Chapter 6 Application and Operation Concepts

Figure 6-58. Typical Setup for the External Source Mode

External Source Mode In-Depth Description

Y ou may use the external source in automatic or manual mode . External source mode phase

locks the analyzer to an external CW signal.

The external source mode works only in CW time sweep .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

External Source A uto . If you press

SYSTEM

INSTRUMENT MODE

EXT SOURCE AUTO the

analyzer turns on the external source auto mode . Y ou should observe the following points

when using this operation mode:

The auto mode has a wider capture range than the manual mode .

The manual mode is faster than the auto mode .

The auto mode searches for the incoming CW signal.

The capture range is typically 10% of the selected CW frequency .

This feature works only in CW time sweep type .

The incoming signal should not have large spurs or sidebands , as the analyzer may phase lock

on a spur or not phase lock at all.

The frequency the instrument has locked onto is shown on the analyzer , and is also available

via HP-IB .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

External Source Manual. If you press

SYSTEM

INSTRUMENT MODE

EXT SOURCE MANUAL the

analyzer activates the external source manual mode . Y ou should observe the following points

when using this operation mode:

The manual mode has a smaller capture range than the auto mode .

The manual mode is much faster than auto mode .

This feature works only in CW time sweep type .

The incoming signal should not have large spurs or sidebands , as the analyzer may phase lock

on a spur or not phase lock at all.

6-110 Application and Operation Concepts

·

Page 378 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 378 · click to enlarge
Page 378

The frequency of the incoming signal should be within 0.5 to +5.0 MHz of the selected

frequency or the analyzer will not be able to phase lock to it.

CW Frequency Range in External Source Mode . 300 kHz to 3 GHz (6 GHz for Option 006)

Compatible Sweep Types . The external source mode will only function in CW time sweep .

If the instrument is in any other sweep type when external source is activated, the warning

message CHANGED TO CW TIME MODE will appear on the display .

External Source Requirements . The external source mode has spectral purity and power

input requirements , which are described in Chapter 7, \Speci cations and Measurement

Uncertainties ."

Input ChannelR

Capture Range . In either automatic or manual mode , you can enter the frequency of the

external CW signal using the

NNNNNNNNNNNNNNNNNNNNNNN

CW FREQ softkey (located under the Stimulus

key). The

MENU

actual signal must be within a certain frequency capture range as shown in T able 6-7.

T able 6-7. External Source Capture Ranges

Mode CW Frequency Capture Range

A utomatic 50 MHz 5 MHz of nominal CW frequency

> 50 MHz 10% of nominal CW frequency

Manual All 0.5 to +5 MHz of nominal CW frequency

If the incoming signal is not within the capture range , the analyzer will not phase lock

correctly .

Locking onto a signal with a frequency modulation component. Although the analyzer may

phase-lock onto a signal that has FM, it may not accurately show the signal's amplitude . The

accuracy of such measurements depends greatly on the IF bandwidth you choose . Use the

widest IF bandwidth available (3 kHz) if this problem occurs .

If you press

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SYSTEM

INSTRUMENT MODE

TUNED RECEIVER the analyzer receiver operates

independently of any signal source .

The following features and limitations apply to the tuned receiver mode:

It is a fully synthesized receiver; it does not phase-lock to any source .

It functions in all sweep types .

It requires a synthesized CW source whose timebase is input to the analyzer's external

frequency reference .

F or more information on using the tuned receiver mode , refer to Chapter 2, \Making

Measurements ."

Application and Operation Concepts 6-111

Tuned Receiver Mode ·

Page 379 of the Agilent Technologies 8753D user manual — SYSTEM
Page 379 · click to enlarge
Page 379

Frequency O set Menu

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

If you press

SYSTEM

INSTRUMENT MODE

FREQ OFFS MENU , the analyzer allows phase-locked

operation with a frequency o set between the internal source and receiver . This feature is

used in swept RF mixer measurements and has an upper frequency limit equal to that of the

analyzer being used.

This feature allows you to set the RF source to a xed o set frequency above or below the

receiver (as required in a mixer test, using a swept RF/IF and xed LO). Then you can input a

signal to a device over one frequency range and view its response over a di erent frequency

range . The maximum delay between the RF source and the R input is 0.3 microseconds . The

analyzer will show a signal that is a composite of the desired RF signal, image response , and

spurious signals .

Y ou can use the frequency o set in any sweep type in network analyzer mode . The two

user-de ned variables in this mode are receiver frequency and o set frequency (LO). The

analyzer automatically sets the source frequency equal to IF + LO or IF LO .

Mixer measurements and frequency o set mode applications are explained in application note

8753-2, \RF Component Measurements Mixer Measurements using the HP 8753B Network

Analyzer ," HP part number 5956-4362. This application note was written for the HP 8753B but

also applies to the HP 8753D . Also see product note 8753-2A, HP part number 5952-2771.

Frequency o set mode is useful for the following types of measurements on frequency-

translating device:

amplitude and phase tracking

Typical T est Setup

Figure 6-59 shows a typical test setup using frequency o set mode . Instructions are provided

in Chapter 3, \Making Mixer Measurements ." The attenuators shown reduce mismatch

uncertainties . The low pass lter keeps unwanted mixing products out of the sampler .

6-112 Application and Operation Concepts

Show 4 more labels / short text on this page
· Primary Applications · conversion loss · conversion compression
Page 380 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 380 · click to enlarge
Page 380
Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 380)
Fig. 175 · Chapter 6 Application and Operation Concepts

Figure 6-59. Typical T est Setup for a Frequency O set Measurement

Frequency O set In-Depth Description

The source and receiver operate at two di erent frequencies in frequency o set operation.

The di erence between the source and receiver frequencies is the LO frequency that you

specify .

The two user-de ned variables in frequency o set are the receiver frequency , and the o set

(LO) frequency . The source frequency is automatically set by the instrument and equals

receiver frequency IF + LO or IF LO .

The Receiver Frequency . Y ou can choose a CW value or start and stop values for the receiver

frequency . The stimulus values , which appear on the analyzer display , will a ect only the

receiver .

The O set Frequency (LO). This frequency value is the di erence between the source and

receiver frequencies .

The analyzer's source locks to the receiver the LO frequency , regardless of

the o set value you selected.

Once the source is phase-locked and sweeping, the analyzer's source frequency

is not known precisely . As the LO frequency changes , the source tracks it to

maintain the receiver start/stop or CW frequency that you requested.

Frequency Hierarchy . The source frequency can be greater than or less than the LO

frequency . That is , the analyzer can measure either the lower or upper of the two IF mixing

products when it is in the frequency o set mode .

Frequency Ranges . Receiver frequency range: 300 kHz to 3 GHz (or 6 GHz with Option 006)

Compatible Instrument Modes and Sweep Types . Frequency o set is compatible with all

sweep types in the network analyzer mode .

Application and Operation Concepts 6-113

Page 381 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 381 · click to enlarge
Page 381

Receiver and Source Requirements . Refer to Chapter 7, \Speci cations and Measurement

Uncertainties ."

IF Input: R always; and port 1 or port 2 for a ratio measurement.

Display Annotations . The analyzer shows the annotation ofs when the frequency o set mode

is on. The annotation of? indicates that the source frequency is approximately 10 MHz away

from the sum of the IF and LO frequencies that you requested. This is most likely caused by

the LO frequency being outside the 1 to +5 MHz accuracy requirement.

Error Message . If you connect your test device before you switch on the frequency o set

function, the error message PHASE LOCK CAL FAILED appears on the screen. This is normal, and

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

will disappear when you press the

FREQ OFFS on OFF softkey .

Spurious Signal P assband Frequencies . Unwanted mixing products (or spurious LO signals)

at speci c frequencies can cause measurement inaccuracy , because of the characteristics of

a sampler . These speci c frequencies can be calculated. Y ou can reduce unwanted mixing

products going to the sampler by inserting a low pass lter at your test device's IF output.

6-114 Application and Operation Concepts

Page 382 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 382 · click to enlarge
Page 382

Harmonic Operation (Option 002 only)

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

The analyzer's harmonic menu can be accessed by pressing

SYSTEM

HARMONIC MEAS .

The harmonic measurement mode allows you to measure the second or third harmonic as the

analyzer's source sweeps fundamental frequencies above 16 MHz. The analyzer can make

harmonic measurements in any sweep type .

Typical T est Setup

Figure 6-60. Typical Harmonic Mode T est Setup

Y ou can view the second or third harmonic alone by using only one of the analyzer's two

channels .

T o make the following types of measurements , uncouple channels 1 and 2, and switch on dual

channel.

The analyzer measures the fundamental on one channel while measuring the second or third

harmonic on the other channel.

The analyzer measures the second harmonic on one channel while measuring the third

harmonic on the other channel.

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

COUPLE PWR ON off feature , the analyzer measures the fundamental on channel 1

while measuring the second or third harmonic in dBc on channel 2.

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

COUPLE PWR ON off feature , the analyzer couples power between channels 1

and 2. This is useful when you are using the D2/D1 to D2 feature because you can change

fundamental power and see the resultant change in the harmonic power .

The analyzer shows the fundamental frequency value on the display . However , a marker in the

active entry area shows the harmonic frequency in addition to the fundamental. If you use

the harmonic mode , the annotation H=2 or H=3 appears on the left-hand side of the display .

The measured harmonic cannot not exceed the frequency limitations of the network analyzer's

receiver .

Coupling P ower Between Channels 1 and 2

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

COUPLE PWR ON off is intended to be used with the

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

D2/D1 toD2 on OFF softkey . Y ou

can use the D2/D1 to D2 function in harmonic measurements , where the analyzer shows

the fundamental on channel 1 and the harmonic on channel 2. D2/D1 to D2 ratios the two ,

Application and Operation Concepts 6-115

Show 5 more labels / short text on this page
· Single-Channel Operation · Dual-Channel Operation · Using the · Using the
Page 383 of the Agilent Technologies 8753D user manual — HP 8753D HP 8753D
Page 383 · click to enlarge
Page 383

showing the fundamental and the relative power of the measured harmonic in dBc . Y ou must

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

uncouple channels 1 and 2 for this measurement, using the

COUPLED CHAN ON off softkey set

to OFF to allow alternating sweeps .

After uncoupling channels 1 and 2, you may want to change the fundamental power and see

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

the resultant change in relative harmonic power (in dBc).

COUPLE PWR ON off allows you to

change the power of both channels simultaneously , even though they are uncoupled in all other

respects .

The frequency range is determined by the upper frequency range of the instrument or system

(3 or 6 GHz) and by the harmonic being displayed. The 6 GHz operation requires an HP 8753D

Option 006. T able 6-8 shows the highest fundamental frequency for maximum frequency and

harmonic mode .

T able 6-8. Maximum Fundamental Frequency using Harmonic Mode

HP 8753D HP 8753D

2nd Harmonic 1.5 GHz 3 GHz

3rd Harmonic 1.0 GHz 2.0 GHz

A ccuracy and input power

Refer to Chapter 7, \Speci cations and Measurement Uncertainties ." The maximum

recommended input power and maximum recommended source power are related

speci cations .

Using power levels greater than the recommended values , you may cause undesired harmonics

in the source and receiver . The recommended power levels ensure that these harmonics are

less than 45 dBc . Use test port power to limit the input power to your test device .

6-116 Application and Operation Concepts

Show 5 more labels / short text on this page
Frequency Range · Harmonic · Maximum Fundamental Frequency · Measured · (Option 006)
Page 384 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 384 · click to enlarge
Page 384

Time Domain Operation (Option 010)

With Option 010, the analyzer can transform frequency domain data to the time domain or

time domain data to the frequency domain.

In normal operation, the analyzer measures the characteristics of a test device as a function

of frequency . Using a mathematical technique (the inverse F ourier transform), the analyzer

transforms frequency domain information into the time domain, with time as the horizontal

display axis . Response values (measured on the vertical axis) now appear separated in time

or distance , providing valuable insight into the behavior of the test device beyond simple

frequency characteristics .

An HP 8753D can be ordered with Option 010, or the option can be added at a

later date using the HP 85019B time domain retro t kit.

The transform used by the analyzer resembles time domain re ectometry (TDR) measurements .

TDR measurements , however , are made by launching an impulse or step into the test device

and observing the response in time with a receiver similar to an oscilloscope . In contrast, the

analyzer makes swept frequency response measurements , and mathematically transforms the

data into a TDR-like display .

The analyzer's transform menu can be accessed by pressing

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SYSTEM

TRANSFORM MENU . This

menu consists of the following softkeys:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SET FREQ LOW PASS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOW PASS IMPULSE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOW PASS STEP

NNNNNNNNNNNNNNNNNNNNNNNNNN

BANDPASS

NNNNNNNNNNNNNNNNNNNN

WINDOW

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SPECIFY GATE

The analyzer has three frequency-to-time transform modes:

Time domain bandpass mode is designed to measure band-limited devices and is the easiest

mode to use . This mode simulates the time domain response of an impulse input.

Time domain low pass step mode simulates the time domain response of a step input. As in

a traditional TDR measurement, the distance to the discontinuity in the test device , and the

type of discontinuity (resistive , capacitive , inductive) can be determined.

Time domain low pass impulse mode simulates the time domain response of an impulse

input (like the bandpass mode). Both low pass modes yield better time domain resolution for

a given frequency span than does the bandpass mode . In addition, when using the low pass

modes , you can determine the type of discontinuity . However , these modes have certain

limitations that are de ned in \Time domain low pass ," later in this section.

Application and Operation Concepts 6-117

The Transform Menu · · TRANSFORM ON off

Page 385 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 385 · click to enlarge
Page 385

The analyzer has one time-to-frequency transform mode:

F orward transform mode transforms CW signals measured over time into the frequency

domain, to measure the spectral content of a signal. This mode is known as the CW time

mode .

In addition to these transform modes , this section discusses special transform concepts such as

masking, windowing, and gating.

The relationship between the frequency domain response and the time domain response of

the analyzer is de ned by the F ourier transform. Because of this transform, it is possible to

measure , in the frequency domain, the response of a linear test device and mathematically

calculate the inverse F ourier transform of the data to nd the time domain response . The

analyzer's internal computer makes this calculation using the chirp-Z F ourier transform

technique . The resulting measurement is the fully error-corrected time domain re ection or

transmission response of the test device , displayed in near real-time .

Figure 6-61 illustrates the frequency and time domain re ection responses of a test device . The

frequency domain re ection measurement is the composite of all the signals re ected by the

discontinuities present in the test device over the measured frequency range .

In this section, all points of re ection are referred to as discontinuities .

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 385)
Fig. 176 · Chapter 6 Application and Operation Concepts

Figure 6-61. Device Frequency Domain and Time Domain Re ection Responses

The time domain measurement shows the e ect of each discontinuity as a function of time

(or distance), and shows that the test device response consists of three separate impedance

changes . The second discontinuity has a re ection coecient magnitude of 0.035 (i.e . 3.5% of

the incident signal is re ected). Marker 1 on the time domain trace shows the elapsed time

from the reference plane (where the calibration standards are connected) to the discontinuity

and back18.2 nanoseconds . The distance shown (5.45 meters) is based on the assumption

that the signal travels at the speed of light. The signal travels slower than the speed of light

in most media (e .g. coax cables). This slower velocity (relative to light) can be compensated

for by adjusting the analyzer relative velocity factor . This procedure is described later in this

section under \Time domain bandpass ."

6-118 Application and Operation Concepts

General Theory

Page 386 of the Agilent Technologies 8753D user manual — CAL
Page 386 · click to enlarge
Page 386

This mode is called bandpass because it works with band-limited devices . Traditional TDR

requires that the test device be able to operate down to dc . Using bandpass mode , there are no

restrictions on the measurement frequency range . Bandpass mode characterizes the test device

impulse response .

A djusting the Relative V elocity F actor

A marker provides both the time (x2) and the electrical length (x2) to a discontinuity . T o

determine the physical length, rather than the electrical length, change the velocity factor to

that of the medium under test:

NNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

1. Press

CAL

MORE

VELOCITY FACTOR .

2. Enter a velocity factor between 0 and 1.0 (1.0 corresponds to the speed of light in a

vacuum). Most cables have a velocity factor of 0.66 (polyethylene dielectrics) or 0.70 (te on

dielectrics).

T o cause the markers to read the actual one-way distance to a discontinuity ,

rather than the two-way distance , enter one-half the actual velocity factor .

Re ection Measurements Using Bandpass Mode

The bandpass mode can transform re ection measurements to the time domain. Figure 6-62a

shows a typical frequency response re ection measurement of two sections of cable .

Figure 6-62b shows the same two sections of cable in the time domain using the bandpass

mode .

Application and Operation Concepts 6-119

Time Domain Bandpass ·

Page 387 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 387 · click to enlarge
Page 387
Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 387)
Fig. 177 · Chapter 6 Application and Operation Concepts

Figure 6-62. A Re ection Measurement of Two Cables

The ripples in re ection coecient versus frequency in the frequency domain measurement are

caused by the re ections at each connector \beating" against each other .

One at a time , loosen the connectors at each end of the cable and observe the response in both

the frequency domain and the time domain. The frequency domain ripples increase as each

connector is loosened, corresponding to a larger re ection adding in and out of phase with

the other re ections . The time domain responses increase as you loosen the connector that

corresponds to each response .

Interpreting the bandpass re ection response horizontal axis . In bandpass re ection

measurements , the horizontal axis represents the time it takes for an impulse launched at the

test port to reach a discontinuity and return to the test port (the two-way travel time). In

Figure 6-61, each connector is a discontinuity .

Interpreting the bandpass re ection response vertical axis . The quantity displayed on the

vertical axis depends on the selected format. The common formats are listed in T able 6-9. The

default format is LOG MA G (logarithmic magnitude), which displays the return loss in decibels

(dB). LIN MA G (linear magnitude) is a format that displays the response as re ection coecient

(). This can be thought of as an average re ection coecient of the discontinuity over the

frequency range of the measurement. Use the REAL format only in low pass mode .

6-120 Application and Operation Concepts

Page 388 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 388 · click to enlarge
Page 388

T able 6-9. Time Domain Re ection F ormats

F ormat P arameter

LIN MA G Re ection Coecient (unitless) (0

REAL Re ection Coecient (unitless) (1

LOG MA G Return Loss (dB)

SWR Standing W ave Ratio (unitless)

Transmission Measurements Using Bandpass Mode

The bandpass mode can also transform transmission measurements to the time domain. F or

example , this mode can provide information about a surface acoustic wave (SA W) lter that

is not apparent in the frequency domain. Figure 6-63 illustrates a time domain bandpass

measurement of a 321 MHz SA W lter .

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 388)
Fig. 178 · Chapter 6 Application and Operation Concepts

Figure 6-63. Transmission Measurement in Time Domain Bandpass Mode

Interpreting the bandpass transmission response horizontal axis . In time domain

transmission measurements , the horizontal axis is displayed in units of time . The time axis

indicates the propagation delay through the device . Note that in time domain transmission

measurements , the value displayed is the actual delay (not x2). The marker provides the

propagation delay in both time and distance .

Marker 2 in Figure 6-63 a indicates the main path response through the test device , which

has a propagation delay of 655.6 ns , or about 196.5 meters in electrical length. Marker 4 in

Figure 6-63b indicates the triple-travel path response at 1.91 s , or about 573.5 meters . The

response at marker 1 (at 0 seconds) is an RF feedthru leakage path. In addition to the triple

travel path response , there are several other multi-path responses through the test device ,

which are inherent in the design of a SA W lter .

Interpreting the bandpass transmission response vertical axis . In the log magnitude

format, the vertical axis displays the transmission loss or gain in dB; in the linear magnitude

format it displays the transmission coecient ( ). Think of this as an average of the

transmission response over the measurement frequency range .

Application and Operation Concepts 6-121

Page 389 of the Agilent Technologies 8753D user manual — N
Page 389 · click to enlarge
Page 389

Time domain low pass

This mode is used to simulate a traditional time domain re ectometry (TDR) measurement. It

provides information to determine the type of discontinuity (resistive , capacitive , or inductive)

that is present. Low pass provides the best resolution for a given bandwidth in the frequency

domain. It may be used to give either the step or impulse response of the test device .

The low pass mode is less general-purpose than the bandpass mode because it places strict

limitations on the measurement frequency range . The low pass mode requires that the

frequency domain data points are harmonically related from dc to the stop frequency . That is ,

stop = n x start, where n = number of points . F or example , with a start frequency of 30 kHz

and 101 points , the stop frequency would be 3.03 MHz. Since the analyzer frequency range

starts at 30 kHz, the dc frequency response is extrapolated from the lower frequency data. The

requirement to pass dc is the same limitation that exists for traditional TDR.

Setting frequency range for time domain low pass

Before a low pass measurement is made , the measurement frequency range must meet the

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

(stop = n x start) requirement described above . The

SET FREQ LOW PASS softkey performs

this function automaticallythe stop frequency is set close to the entered stop frequency , and

the start frequency is set equal to stop/n.

If the low end of the measurement frequency range is critical, it is best to calculate

approximate values for the start and stop frequencies before pressing

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SET FREQ LOW PASS

and calibrating. This avoids distortion of the measurement results . T o see an example ,

select the preset values of 201 points and a 300 kHz to 3 GHz frequency range . Now press

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SET FREQ LOW PASS and observe the change in frequency values . The stop frequency changes

to 2.999 GHz, and the start frequency changes to 14.925 MHz. This would cause a distortion of

measurement results for frequencies from 300 kHz to 14.925 MHz.

If the start and stop frequencies do not conform to the low pass requirement

before a low pass mode (step or impulse) is selected and transform is turned

on, the analyzer resets the start and stop frequencies . If error correction is on

when the frequency range is changed, this turns it o .

T able 6-10. Minimum Frequency Ranges for Time Domain Low P ass

Number of P oints Minimum Frequency Range

3 30 kHz to 0.09 MHz

11 30 kHz to 0.33 MHz

26 30 kHz to 0.78 MHz

51 30 kHz to 1.53 MHz

101 30 kHz to 3.03 MHz

201 30 kHz to 6.03 MHz

401 30 kHz to 12.03 MHz

801 30 kHz to 24.03 MHz

1601 30 kHz to 48.03 MHz

6-122 Application and Operation Concepts

Page 390 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 390 · click to enlarge
Page 390

Minimum allowable stop frequencies . The lowest analyzer measurement frequency is 30

kHz, therefore for each value of n there is a minimum allowable stop frequency that can

be used. That is , the minimum stop frequency =n x 30 kHz. T able 6-10 lists the minimum

frequency range that can be used for each value of n for low pass time domain measurements .

Re ection Measurements In Time Domain Low P ass

Figure 6-64 shows the time domain response of an unterminated cable in both the low-pass step

and low-pass impulse modes .

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 390)
Fig. 179 · Chapter 6 Application and Operation Concepts

Figure 6-64.

Time Domain Low P ass Measurements of an Unterminated Cable

Interpreting the low pass response horizontal axis . The low pass measurement horizontal

axis is the two-way travel time to the discontinuity (as in the bandpass mode). The marker

displays both the two-way time and the electrical length along the trace . T o determine the

actual physical length, enter the appropriate velocity factor as described earlier in this section

under \Time domain bandpass ."

Interpreting the low pass response vertical axis . The vertical axis depends on the chosen

format. In the low pass mode , the frequency domain data is taken at harmonically related

frequencies and extrapolated to dc . Because this results in the inverse F ourier transform

having only a real part (the imaginary part is zero), the most useful low pass step mode format

in this application is the real format. It displays the response in re ection coecient units .

This mode is similar to the traditional TDR response , which displays the re ected signal in a

real format (volts) versus time (or distance) on the horizontal axis .

The real format can also be used in the low pass impulse mode , but for the best dynamic range

for simultaneously viewing large and small discontinuities , use the log magnitude format.

F ault Location Measurements Using Low P ass

As described, the low pass mode can simulate the TDR response of the test device . This

response contains information useful in determining the type of discontinuity present.

Figure 6-65 illustrates the low pass responses of known discontinuities . Each circuit element

was simulated to show the corresponding low pass time domain S

response waveform.

The low pass mode gives the test device response either to a step or to an impulse stimulus .

Mathematically , the low pass impulse stimulus is the derivative of the step stimulus .

Application and Operation Concepts 6-123

Page 391 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 391 · click to enlarge
Page 391
Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 391)
Fig. 180 · Chapter 6 Application and Operation Concepts

Figure 6-65. Simulated Low P ass Step and Impulse Response W aveforms (Real F ormat)

Figure 6-66 shows example cables with discontinuities (faults) using the low pass step mode

with the real format.

6-124 Application and Operation Concepts

Page 392 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 392 · click to enlarge
Page 392
Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 392)
Fig. 181 · Chapter 6 Application and Operation Concepts

Figure 6-66. Low P ass Step Measurements of Common Cable F aults (Real F ormat)

Transmission Measurements In Time Domain Low P ass

Measuring small signal transient response using low pass step . Use the low pass mode to

analyze the test device's small signal transient response . The transmission response of a device

to a step input is often measured at lower frequencies , using a function generator (to provide

the step to the test device) and a sampling oscilloscope (to analyze the test device output

response). The low pass step mode extends the frequency range of this type of measurement to

3 GHz (6 GHz with an analyzer Option 006).

The step input shown in Figure 6-67 is the inverse F ourier transform of the frequency domain

response of a thru measured at calibration. The step rise time is proportional to the highest

frequency in the frequency domain sweep; the higher the frequency , the faster the rise time .

The frequency sweep in Figure 6-67 is from 10 MHz to 1 GHz.

Figure 6-67 also illustrates the time domain low pass response of an ampli er under test. The

average group delay over the measurement frequency range is the di erence in time between

the step and the ampli er response . This time domain response simulates an oscilloscope

measurement of the ampli er's small signal transient response . Note the ringing in the

ampli er response that indicates an under-damped design.

Application and Operation Concepts 6-125

Page 393 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 393 · click to enlarge
Page 393
Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 393)
Fig. 182 · Chapter 6 Application and Operation Concepts

Figure 6-67.

Time Domain Low P ass Measurement of an Ampli er Small Signal

Interpreting the low pass step transmission response horizontal axis . The low pass

transmission measurement horizontal axis displays the average transit time through the

test device over the frequency range used in the measurement. The response of the thru

connection used in the calibration is a step that reaches 50% unit height at approximately

time = 0. The rise time is determined by the highest frequency used in the frequency domain

measurement. The step is a unit high step , which indicates no loss for the thru calibration.

When a device is inserted, the time axis indicates the propagation delay or electrical length

of the device . The markers read the electrical delay in both time and distance . The distance

can be scaled by an appropriate velocity factor as described earlier in this section under \Time

domain bandpass ."

Interpreting the low pass step transmission response vertical axis . In the real format,

the vertical axis displays the transmission response in real units (for example , volts). F or the

ampli er example in Figure 6-67, if the ampli er input is a step of 1 volt, the output,

2.4 nanoseconds after the step (indicated by marker 1), is 5.84 volts .

In the log magnitude format, the ampli er gain is the steady state value displayed after the

initial transients die out.

Measuring separate transmission paths through the test device using low pass impulse

mode . The low pass impulse mode can be used to identify di erent transmission paths through

a test device that has a response at frequencies down to dc (or at least has a predictable

response , above the noise oor , below 30 kHz). F or example , use the low pass impulse mode to

measure the relative transmission times through a multi-path device such as a power divider .

Another example is to measure the pulse dispersion through a broadband transmission line ,

such as a ber optic cable . Both examples are illustrated in Figure 6-68 . The horizontal and

vertical axes can be interpreted as already described in this section under \Time Domain

Bandpass".

6-126 Application and Operation Concepts

Transient Response

Page 394 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 394 · click to enlarge
Page 394
Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 394)
Fig. 183 · Chapter 6 Application and Operation Concepts

Figure 6-68. Transmission Measurements Using Low P ass Impulse Mode

Masking occurs when a discontinuity (fault) closest to the reference plane a ects the response

of each subsequent discontinuity . This happens because the energy re ected from the rst

discontinuity never reaches subsequent discontinuities . F or example , if a transmission line has

two discontinuities that each re ect 50% of the incident voltage , the time domain response

(real format) shows the correct re ection coecient for the rst discontinuity (=0.50).

However , the second discontinuity appears as a 25% re ection (=0.25) because only half the

incident voltage reached the second discontinuity .

This example assumes a loss-less transmission line . Real transmission lines , with

non-zero loss , attenuate signals as a function of the distance from the reference

plane .

As an example of masking due to line loss , consider the time domain response of a 3 dB

attenuator and a short circuit. The impulse response (log magnitude format) of the short circuit

alone is a return loss of 0 dB , as shown in Figure 6-69a. When the short circuit is placed at

the end of the 3 dB attenuator , the return loss is 6 dB , as shown in Figure 6-69b . This value

actually represents the forward and return path loss through the attenuator , and illustrates

how a lossy network can a ect the responses that follow it.

Application and Operation Concepts 6-127

Time Domain Concepts · Masking

Page 395 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 395 · click to enlarge
Page 395
Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 395)
Fig. 184 · Chapter 6 Application and Operation Concepts

Figure 6-69. Masking Example

The analyzer provides a windowing feature that makes time domain measurements more useful

for isolating and identifying individual responses . Windowing is needed because of the abrupt

transitions in a frequency domain measurement at the start and stop frequencies . The band

limiting of a frequency domain response causes overshoot and ringing in the time domain

response , and causes a non-windowed impulse stimulus to have a sin(kt)/kt shape , where

k = /frequency span and t = time (see Figure 6-70). This has two e ects that limit the

usefulness of the time domain measurement:

Finite impulse width (or rise time). Finite impulse width limits the ability to resolve

between two closely spaced responses . The e ects of the nite impulse width cannot be

improved without increasing the frequency span of the measurement (see T able 6-11 ).

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 395)
Fig. 185 · Chapter 6 Application and Operation Concepts

Figure 6-70. Impulse Width, Sidelobes , and Windowing

Sidelobes . The impulse sidelobes limit the dynamic range of the time domain measurement

by hiding low-level responses within the sidelobes of higher level responses . The e ects of

sidelobes can be improved by windowing (see T able 6-11).

Windowing improves the dynamic range of a time domain measurement by ltering the

frequency domain data prior to converting it to the time domain, producing an impulse

stimulus that has lower sidelobes . This makes it much easier to see time domain responses that

are very di erent in magnitude . The sidelobe reduction is achieved, however , at the expense of

6-128 Application and Operation Concepts

Windowing

Page 396 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 396 · click to enlarge
Page 396

increased impulse width. The e ect of windowing on the step stimulus (low pass mode only) is

a reduction of overshoot and ringing at the expense of increased rise time .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNN

T o select a window , press

SYSTEM

TRANSFORM MENU

WINDOW . A menu is presented that allows

the selection of three window types (see T able 6-11 ).

T able 6-11. Impulse Width, Sidelobe Level, and Windowing V alues

Minimum 13 dB 0.60/Freq Span 21 dB 0.45/Freq Span

Normal 44 dB 0.98/Freq Span 60 dB 0.99/Freq Span

Maximum 75 dB 1.39/Freq Span 70 dB 1.48/Freq Span

impulse width. The bandpass impulse sidelobe levels are the same as low pass impulse sidelobe levels .

Choose one of the three window shapes listed in T able 6-11 . Or you can use the knob to select

any windowing pulse width (or rise time for a step stimulus) between the softkey values . The

time domain stimulus sidelobe levels depend only on the window selected.

N

NNNNNNNNNNNNNNNNNNNNNN

MINIMUM is essentially no window . Consequently , it gives the highest

sidelobes .

NNNNNNNNNNNNNNNNNNNN

NORMAL (the preset mode) gives reduced sidelobes and is the mode

most often used.

NNNNNNNNNNNNNNNNNNNNNNN

MAXIMUM window gives the minimum sidelobes , providing the

greatest dynamic range .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

USE MEMORY on OFF remembers a user-speci ed window pulse width (or step

rise time) di erent from the standard window values .

A window is activated only for viewing a time domain response , and does not a ect a

displayed frequency domain response . Figure 6-71 shows the typical e ects of windowing on

the time domain response of a short circuit re ection measurement.

Application and Operation Concepts 6-129

Show 15 more labels / short text on this page
· Window · Impulse · Low P ass · Step · Step · Type · Sidelobe · Impulse · Sidelobe · Rise Time · Level · Width (50%) · Level · (10 - 90%)
Page 397 of the Agilent Technologies 8753D user manual —  F
Page 397 · click to enlarge
Page 397
Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 397)
Fig. 186 · Chapter 6 Application and Operation Concepts

Figure 6-71. The E ects of Windowing on the Time Domain Responses of a Short Circuit

In the time domain, range is de ned as the length in time that a measurement can be made

without encountering a repetition of the response , called aliasing. A time domain response

repeats at regular intervals because the frequency domain data is taken at discrete frequency

points , rather than continuously over the frequency band.

M easur ement r ang e =

F

where F is the spacing between frequency data points

(number of points 1)

M easur ement r ang e =

f r eq uency span(H z )

example:

M easur ement = 201 points

1 M H z to 2: 001 GH z

(number of points 1)

R ang e =

F

f r eq uency span

= 100 10

E l ectr ical l eng th = r ang e the speed of l ig ht (3 10

= (100 10

s) (3 10

= 30 meter s

In this example , the range is 100 ns , or 30 meters electrical length. T o prevent the time domain

responses from overlapping, the test device must be 30 meters or less in electrical length for a

transmission measurement (15 meters for a re ection measurement). The analyzer limits the

stop time to prevent the display of aliased responses .

6-130 Application and Operation Concepts

Show 14 more labels / short text on this page
Range · or · (201 1) · = · (10 10 · ) · or · (2 10 · ) · 9 · seconds · m=s) · 9 · m=s)
Page 398 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 398 · click to enlarge
Page 398

T o increase the time domain measurement range , rst increase the number of points , but

remember that as the number of points increases , the sweep speed decreases . Decreasing the

frequency span also increases range , but reduces resolution.

Two di erent resolution terms are used in the time domain:

Response resolution. Time domain response resolution is de ned as the ability to resolve two

closely-spaced responses , or a measure of how close two responses can be to each other and

still be distinguished from each other . F or responses of equal amplitude , the response resolution

is equal to the 50% (6 dB) impulse width. It is inversely proportional to the measurement

frequency span, and is also a function of the window used in the transform. The approximate

formulas for calculating the 50% impulse width are given in T able 6-11 . F or example , using the

formula for the bandpass mode with a normal windowing function for a 50 MHz to 13.05 GHz

measurement (13.0 GHz span):

0. 98

50% cal cul ated impul se w idth =

13. 0 ( GH z )
= 0151 nanoseconds

E l ectr ical l eng th (in air ) = (0: 151 10

s) (30 10

= 453 centimeter s

With this measurement, two equal responses can be distinguished when they are separated

by at least 4.53 centimeters . In a measurement with a 20 GHz span, two equal responses can

be distinguished when they are separated by at least 2.94 cm. Using the low pass mode (the

low pass frequencies are slightly di erent) with a minimum windowing function, you can

distinguish two equal responses that are about 1.38 centimeters or more apart.

F or re ection measurements , which measure the two-way time to the response , divide the

response resolution by 2. Using the example above , you can distinguish two faults of equal

magnitude provided they are 0.69 centimeters (electrical length) or more apart.

Remember , to determine the physical length, the relative velocity factor of

the transmission medium under test must be entered into the electrical length

equation.

F or example , a cable with a te on dielectric (0.7 relative velocity factor), measured under the

conditions stated above , has a fault location measurement response resolution of

0.45 centimeters . This is the maximum fault location response resolution. F actors such as

reduced frequency span, greater frequency domain data windowing, and a large discontinuity

shadowing the response of a smaller discontinuity , all act to degrade the e ective response

resolution.

Figure 6-72 illustrates the e ects of response resolution. The solid line shows the actual

re ection measurement of two approximately equal discontinuities (the input and output of an

SMA barrel). The dashed line shows the approximate e ect of each discontinuity , if they could

be measured separately .

Application and Operation Concepts 6-131

Show 6 more labels / short text on this page
Resolution · response resolution · range resolution · 2 · 9 · cm=s)
Page 399 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 399 · click to enlarge
Page 399
Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 399)
Fig. 187 · Chapter 6 Application and Operation Concepts

Figure 6-72. Response Resolution

While increasing the frequency span increases the response resolution, keep the following

points in mind:

The time domain response noise oor is directly related to the frequency domain data

noise oor . Because of this , if the frequency domain data points are taken at or below the

measurement noise oor , the time domain measurement noise oor is degraded.

The time domain measurement is an average of the response over the frequency range of

the measurement. If the frequency domain data is measured out-of-band, the time domain

measurement is also the out-of-band response .

Y ou may (with these limitations in mind) choose to use a frequency span that is wider than the

test device bandwidth to achieve better resolution.

Range resolution. Time domain range resolution is de ned as the ability to locate a single

response in time . If only one response is present, range resolution is a measure of how closely

you can pinpoint the peak of that response . The range resolution is equal to the digital

resolution of the display , which is the time domain span divided by the number of points on the

display . T o get the maximum range resolution, center the response on the display and reduce

the time domain span. The range resolution is always much ner than the response resolution

(see Figure 6-73).

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 399)
Fig. 188 · Chapter 6 Application and Operation Concepts

Figure 6-73. Range Resolution of a Single Discontinuity

6-132 Application and Operation Concepts

Page 400 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 400 · click to enlarge
Page 400

Gating provides the exibility of selectively removing time domain responses . The remaining

time domain responses can then be transformed back to the frequency domain. F or re ection

(or fault location) measurements , use this feature to remove the e ects of unwanted

discontinuities in the time domain. Y ou can then view the frequency response of the remaining

discontinuities . In a transmission measurement, you can remove the e ects of multiple

transmission paths .

Figure 6-74a shows the frequency response of an electrical airline and termination.

Figure 6-74b shows the response in the time domain. The discontinuity on the left is due to the

input connector . The discontinuity on the right is due to the termination. W e want to remove

the e ect of the connector so that we can see the frequency response of just the airline and

termination. Figure 6-74c shows the gate applied to the connector discontinuity . Figure 6-74d

shows the frequency response of the airline and termination, with the connector \gated out."

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 400)
Fig. 189 · Chapter 6 Application and Operation Concepts

Figure 6-74. Sequence of Steps in Gating Operation

Setting the gate . Think of a gate as a bandpass lter in the time domain (see Figure 6-75 ).

When the gate is on, responses outside the gate are mathematically removed from the time

domain trace . Enter the gate position as a start and stop time (not frequency) or as a center

and span time . The start and stop times are the bandpass lter 6 dB cuto times . Gates can

have a negative span, in which case the responses inside the gate are mathematically removed.

The gate's start and stop ags de ne the region where gating is on.

Application and Operation Concepts 6-133

Gating

Page 401 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 401 · click to enlarge
Page 401

Figure 6-75. Gate Shape

Selecting gate shape . The four gate shapes available are listed in T able 6-12 . Each gate has a

di erent passband atness , cuto rate , and sidelobe levels .

T able 6-12. Gate Characteristics

Gate Span Minimum 0.10 dB 48 dB 1.4/Freq Span 2.8/Freq Span

Normal 0.01 dB 68 dB 2.8/Freq Span 5.6/Freq Span

Wide 0.01 dB 57 dB 4.4/Freq Span 8.8/Freq Span

Maximum 0.01 dB 70 dB 12.7/Freq Span 25.4/Freq Span

The passband ripple and sidelobe levels are descriptive of the gate shape . The cuto time is the

time between the stop time (6 dB on the lter skirt) and the peak of the rst sidelobe , and

is equal on the left and right side skirts of the lter . Because the minimum gate span has no

passband, it is just twice the cuto time . Always choose a gate span wider than the minimum.

F or most applications , do not be concerned about the minimum gate span, simply use the knob

to position the gate markers around the desired portion of the time domain trace .

Transforming CW Time Measurements Into the Frequency Domain

The analyzer can display the amplitude and phase of CW signals versus time . F or example , use

this mode for measurements such as ampli er gain as a function of warm-up time (i.e . drift).

The analyzer can display the measured parameter (e .g. ampli er gain) for periods of up to

24 hours and then output the data to a digital plotter for hardcopy results .

These \strip chart" plots are actually measurements as a function of time (time is the

independent variable), and the horizontal display axis is scaled in time units . Transforms of

these measurements result in frequency domain data. Such transforms are called forward

transforms because the transform from time to frequency is a forward F ourier transform, and

can be used to measure the spectral content of a CW signal. F or example , when transformed

into the frequency domain, a pure CW signal measured over time appears as a single frequency

spike . The transform into the frequency domain yields a display that looks similar to a

spectrum analyzer display of signal amplitude versus frequency .

6-134 Application and Operation Concepts

Show 10 more labels / short text on this page
Gate · P assband · Sidelobe · Cuto · Minimum · Shape · Ripple · Levels · Time · Gate Span
Page 402 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 402 · click to enlarge
Page 402

F orward Transform Measurements

This is an example of a measurement using the F ourier transform in the forward direction,

from the time domain to the frequency domain (see Figure 6-76):

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 402)
Fig. 190 · Chapter 6 Application and Operation Concepts

Figure 6-76. Ampli er Gain Measurement

Interpreting the forward transform vertical axis . With the log magnitude format selected,

the vertical axis displays dB . This format simulates a spectrum analyzer display of power versus

frequency .

Interpreting the forward transform horizontal axis . In a frequency domain transform of

a CW time measurement, the horizontal axis is measured in units of frequency . The center

frequency is the o set of the CW frequency . F or example , with a center frequency of 0 Hz, the

CW frequency (250 MHz in the example) is in the center of the display . If the center frequency

entered is a positive value , the CW frequency shifts to the right half of the display; a negative

value shifts it to the left half of the display . The span value entered with the transform on is

the total frequency span shown on the display . (Alternatively , the frequency display values can

be entered as start and stop .)

Demodulating the results of the forward transform. The forward transform can separate

the e ects of the CW frequency modulation amplitude and phase components . F or example , if

a test device modulates the transmission response (S

) with a 500 Hz AM signal, you can see

the e ects of that modulation as shown in Figure 6-77. T o simulate this e ect, apply a 500 Hz

sine wave to the analyzer rear panel EXT AM input.

Application and Operation Concepts 6-135

Page 403 of the Agilent Technologies 8753D user manual — PHASE
Page 403 · click to enlarge
Page 403
Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 403)
Fig. 191 · Chapter 6 Application and Operation Concepts

Figure 6-77. Combined E ects of Amplitude and Phase Modulation

Using the demodulation capabilities of the analyzer , it is possible to view the amplitude or

the phase component of the modulation separately . The window menu includes the following

softkeys to control the demodulation feature:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DEMOD: OFF

is the normal preset state , in which both the amplitude and phase components

of any test device modulation appear on the display .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AMPLITUDE

displays only the amplitude modulation, as illustrated in Figure 6-78a.

NNNNNNNNNNNNNNNNN

displays only the phase modulation, as shown in Figure 6-78b .

PHASE

Agilent Technologies 8753D — PHASE diagram (page 403)
Fig. 192

Figure 6-78.

Separating the Amplitude and Phase Components of T est-Device-Induced Modulation

F orward transform range . In the forward transform (from CW time to the frequency domain),

range is de ned as the frequency span that can be displayed before aliasing occurs , and is

similar to range as de ned for time domain measurements . In the range formula, substitute

time span for frequency span.

Example:

6-136 Application and Operation Concepts

Page 404 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 404 · click to enlarge
Page 404

N umber of points 1

R ang e =

= 1000 H er tz

F or the example given above , a 201 point CW time measurement made over a 200 ms time

span, choose a span of 1 kHz or less on either side of the center frequency (see Figure 6-79 ).

That is , choose a total span of 2 kHz or less .

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 404)
Fig. 193 · Chapter 6 Application and Operation Concepts

Figure 6-79. Range of a F orward Transform Measurement

T o increase the frequency domain measurement range , increase the span. The maximum

range is inversely proportional to the sweep time , therefore it may be necessary to increase

the number of points or decrease the sweep time . Because increasing the number of points

increases the auto sweep time , the maximum range is 2 kHz on either side of the selected CW

time measurement center frequency (4 kHz total span). T o display a total frequency span of

4 kHz, enter the span as 4000 Hz.

Application and Operation Concepts 6-137

Show 5 more labels / short text on this page
time span · 201 1 · = · 200 10 · 3
Page 405 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNN
Page 405 · click to enlarge
Page 405

T est sequencing is an analyzer function that allows you to automate repetitive tasks . Y ou can

create a sequence as you are making a measurement. Then when you want to make that same

measurement again, you can recall the sequence and the analyzer will repeat the previous

keystrokes .

The following is a list of some of the key test sequencing features available on the

HP 8753D network analyzer:

Limited decision-making functions increase the versatility of the test sequences you create by

allowing you to jump from one sequence to another .

A

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GOSUB SEQUENCE function that allows you to call other sequences as sub-routines .

Y ou can create , title , save , and execute up to six sequences .

Y ou can save your sequences to a disk using the internal disk drive .

Y ou can use the parallel port as a general purpose input/output (GPIO) bus to read ve

TTL input bits in a decision making function, and send eight TTL output bits to control a

peripheral.

Product note 8753-3 \RF Component Measurements Applications of the

T est Sequence Function" provides practical applications examples for test

sequencing. This note was written for the HP 8753B but also applies to the

HP 8753D .

F eatures That Operate Di erently When Executed In a Sequence

The analyzer does not allow you to use the following keys in a sequence:

PRESET

Commands That Sequencing Completes Before the Next Sequence Command

The analyzer completes all operations related to the following commands before continuing

with another sequence command:

recall or save (internal or external)

copy list values and operating parameters

CHAN1, CHAN2, W ait 0*

NNNNNNNNNNNNNNNNNNNN

*W ait 0 is the special sequencing function

WAIT x with a zero entered for the delay value .

6-138 Application and Operation Concepts

Show 17 more labels / short text on this page
T est Sequencing · In-Depth Sequencing Information · · * · keys · + · · key · · key · Begins · single sweep · number of groups · auto scale · marker search · marker function · data ! memory
Page 406 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNN
Page 406 · click to enlarge
Page 406

Commands That Require a Clean Sweep

Many front panel commands disrupt the sweep in progress . F or example , changing the channel

or measurement type . When the analyzer does execute a disruptive command in a sequence ,

some instrument functions are inhibited until a complete sweep is taken. This applies to the

following functions:

F orward Stepping In Edit Mode

In the sequence modify mode , you can step through the selected sequence list, where the

analyzer executes each step .

A title may contain non-printable or special ASCII characters if you download it from an

external controller . A non-printable character is represented on the display as .

A sequence may contain up to 2 kbytes of instructions . Typically , this is around 200 sequence

command lines . T o estimate a sequence's size (in kbytes), use the following guidelines .

Type of Command Size in Bytes

Title string character 1

A ctive entry command 1 per digit

Embedding the V alue of the Loop Counter In a Title

Y ou can append a sequentially increasing or decreasing numeric value to the title of stored data

by placing a

NNNNNNNNNNNNNN

MORE

NNNNNNNNNNNNNNNNN

TITLE

NNNNNNNNNNNNNN

MORE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOOP COUNTER command after the title string. (Y ou

DISPLA Y

must limit the title to three characters if you will use it as a disk le name . The three-character

title and ve-digit loop counter number reach the eight-character limit for disk le names .)

This feature is useful in data logging applications .

Y ou can de ne a sequence to run automatically when you apply power to the analyzer . T o

make an autostarting sequence , create a sequence in position six and title it \A UTO". T o stop

an autostarting sequence , press

. T o stop an autostarting sequence from engaging at

LOCAL

power on, you must clear it from memory or rename it.

The instrument's parallel port can be used in two di erent modes . By pressing

LOCAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNN

PARALLEL [ ] softkey , you can select either the

[COPY] mode or the

NNNNNNNNNNNNNNNNNNNN

[GPIO] mode .

The GPIO mode switches the parallel port into a \general purpose input/output" port.

In this mode , the port can be connected to test xtures , power supplies , and other peripheral

equipment that the analyzer can interact with through test sequencing.

Application and Operation Concepts 6-139

Show 10 more labels / short text on this page
autoscale · data ! memory · Titles · Sequence Size · Typical command 2 · · A utostarting Sequences · The GPIO Mode · and · then toggling the
Page 407 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 407 · click to enlarge
Page 407

SEQ

key accesses the Sequencing menu. This menu leads to a series of menus

that allow you to create and control sequences .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GOSUB SEQUENCE softkey , located in the Sequencing menu, activates a feature that allows

the sequence to branch o to another sequence , then return to the original sequence . F or

example , you could perform an ampli er measurement in the following manner:

1. Create sequence 1 for the speci c purpose of performing the gain measurement and printing

the results . This sequence will act as a sub-routine .

2. Create sequence 2 to set up a series of di erent input power levels for the ampli er gain

measurements . In-between each power level setting, call sequence 1 as a sub-routine by

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GOSUB SEQUENCE

SEQUENCE 1 . Now , sequence 2 will print the measurement

results for each input power level applied to the ampli er .

NNNNNNNNNNNNNNNNNNNNNNN

This menu can be accessed by pressing

TTL I/O in the Sequencing menu.

TTL Output for Controlling P eripherals

Eight TTL compatible output lines can be used for controlling equipment connected to the

parallel port. By pressing

SEQ

NNNNNNNNNNNNNNNNNNNNNNN

TTL I/O you will access the softkeys (listed below) that control

the individual output bits . Refer to Figure 6-80 for output bus pin locations .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PARALLEL OUT ALL lets you input a number (0 to 255) in base 10 and outputs it to the bus

as binary .

NNNNNNNNNNNNNNNNNNNNNNN

SET BIT lets you set a single bit (0 - 7) to high on the output bus .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CLEAR BIT lets you set a single bit (0 - 7) to low on the output bus .

TTL Input Decision Making

Five TTL compatible input lines can be used for decision making in test sequencing. F or

example , if a test xture is connected to the parallel port and has a micro switch that needs to

be activated in order to proceed with a measurement, you can construct your test sequence

so that it checks the TTL state of the input line corresponding to the switch. Depending on

whether the line is high or low , you can jump to another sequence . T o access these decision

making functions , press

NNNNNNNNNNNNNNNNNNNNNNN

TTL I/O . Refer to Figure 6-80 for input bus pin locations .

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PARALL IN BIT NUMBER lets you select the single bit (0 - 4) that the sequence will be

looking for .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PARALL IN IF BIT H lets you jump to another sequence if the single input bit you selected

is in a high state .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PARALL IN IF BIT L lets you jump to another sequence if the single input bit you selected

is in a low state .

Pin assignments:

pin 1 is the data strobe

pin 16 selects the printer

6-140 Application and Operation Concepts

Show 8 more labels / short text on this page
The Sequencing Menu · Pressing the · Gosub Sequence Command · The · pressing · TTL I/O Menu · ·
Page 408 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 408 · click to enlarge
Page 408

pin 17 resets the printer

pins 18-25 are ground

Electrical speci cations for TTL high:

volts(H) = 2.7 volts (V)

current = 20 microamps (A)

Electrical speci cations for TTL low:

volts(L) = 0.4 volts (V)

current = 0.2 milliamps (mA)

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 408)
Fig. 194 · Chapter 6 Application and Operation Concepts

Figure 6-80. P arallel P ort Input and Output Bus Pin Locations in GPIO Mode

Application and Operation Concepts 6-141

Page 409 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 409 · click to enlarge
Page 409

NNNNNNNNNNNNNNNNNNNNNNN

TTL OUT softkey provides access to the TTL out menu. This menu allows you to choose

between the following output parameters of the TTL output signal:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TTL OUT HIGH

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TTL OUT LOW

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

END SWEEP HIGH PULSE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

END SWEEP LOW PULSE

The TTL output signals are sent to the sequencing BNC rear panel output.

Sequencing Special Functions Menu

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

This menu is accessed by pressing the

SPECIAL FUNCTIONS softkey in the Sequencing menu.

Sequence Decision Making Menu

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

This menu is accessed by pressing the

DECISION MAKING softkey in the Sequencing Special

Functions menu.

Decision making functions are explained in more detail below . These functions check a

condition and jump to a speci ed sequence if the condition is true . The sequence called must

be in memory . A sequence call is a one-way jump . A sequence can jump to itself, or to any of

the other ve sequences currently in memory . Use of these features is explained under the

speci c softkey descriptions .

Decision making functions jump to a softkey location, not to a speci c sequence

Limit test, loop counter , and do sequence commands jump to any sequence residing in the

speci ed sequence position (1 through 6). These commands do not jump to a speci c sequence

title . Whatever sequence is in the selected softkey position will run when these commands are

executed.

Having a sequence jump to itself

A decision making command can jump to the sequence it is in. When this occurs , the sequence

starts over and all commands in the sequence are repeated. This is used a great deal in

conjunction with loop counter commands . See the loop counter description below .

TTL input decision making

TTL input from a peripheral connected to the parallel port (in the GPIO mode) can be used in a

decision making function. Refer to \The GPIO Mode " earlier in this section.

Limit test decision making

A sequence can jump to another sequence or start over depending on the result of a limit

test. When entered into a sequence , the

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

IF LIMIT TEST PASS and

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

IF LIMIT TEST FAIL

commands require you to enter the destination sequence .

6-142 Application and Operation Concepts

Show 4 more labels / short text on this page
TTL Out Menu · The · Decision Making Functions · title
Page 410 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNN
Page 410 · click to enlarge
Page 410

Loop counter decision making

The analyzer has a numeric register called a loop counter . The value of this register can be set

by a sequence , and it can be incremented or decremented each time a sequence repeats itself.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

The decision making commands

IF LOOP COUNTER = 0 and

IF LOOP COUNTER 0 jump

to another sequence if the stated condition is true . When entered into the sequence , these

commands require you to enter the destination sequence . Either command can jump to another

sequence , or restart the current sequence .

As explained earlier in \Embedding the V alue of The Loop Counter in The Title ," the loop

counter value can be appended to a title . This allows customized titles for data printouts or for

data les saved to disk.

Naming Files Generated by a Sequence

The analyzer can automatically increment the name of a le that is generated by a sequence

using a loop structure .

T o access the sequence lename menu, press:

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FILE UTILITIES

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQUENCE FILENAMING

This menu presents two choices:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FILE NAME FILE0 supplies a name for the saved state and or data le . This also brings up

the Title File Menu.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOT NAME PLOTFILE supplies a name for the plot le generated by a plot-to-disk command.

This also brings up the Title File Menu.

The above keys show the current lename in the 2nd line of the softkey .

When titling a le for use in a loop function, you are restricted to only 2 characters in the

lename due to the 6 character length of the loop counter keyword \[LOOP]." When the le is

actually written, the [LOOP] keyword is expanded to only 5 ASCII characters (digits), resulting

in a 7 character lename .

After entering the 2 character lename , press:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNN

LOOP COUNTER

DONE

The analyzer allows you to use HP-GL (Hewlett-P ackard Graphics Language) to customize

messages or illustrations on the display of the analyzer . T o use HP-GL, the instrument must be

in system controller mode .

HP-GL commands should be entered into a title string using the

NNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNN

DISPLA Y

MORE

character selection menu.

Application and Operation Concepts 6-143

Show 6 more labels / short text on this page
· · HP-GL Considerations · Entering HP-GL Commands · · TITLE and
Page 411 of the Agilent Technologies 8753D user manual — DISPLA Y
Page 411 · click to enlarge
Page 411

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TITLE TO PERIPHERAL sequencing command (in the Sequencing Special Functions

menu) sends the HP-GL command string to the analyzer's HP-GL address . The address of the

analyzer's HP-GL graphics interface is always o set from the instrument's HP-IB address by 1:

If the current instrument address is an even number:

HP-GL address = instrument address +1.

If the current instrument address is an odd number:

HP-GL address = instrument address 1.

Two HP-GL commands require special consideration when used in local operation or in

sequencing. These are explained below:

Plot absolute (HP-GL commandP A)

The syntax for this command is P Ax,y where x and y are screen location coordinates

separated by a comma.

Label (HP-GL commandLB)

The syntax for this command is LB[text][etx]. The label command will print ASCII characters

until the etx command is seen. The etx is the ASCII value 3 (not the ASCII character 3).

The analyzer title function does not have the ASCII value 3, so the instrument allows the LB

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

END OF LABEL command (accessed by pressing

command to be terminated with the

DISPLA Y

NNNNNNNNNNNNNN

MORE

NNNNNNNNNNNNNNNNN

TITLE

NNNNNNNNNNNNNN

MORE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

END OF LABEL ).

Entering Sequences Using HP-IB

Y ou can create a sequence in a computer controller using HP-IB codes and enter it into the

analyzer over HP-IB . This method replaces the keystrokes with HP-IB commands . The following

is a procedure for entering a sequence over HP-IB:

1. Send the HP-IB command NEWSEQx where x is a number from 1 to 6.
2. Send the HP-IB commands for the measurement.
3. T erminate with the HP-IB command DONM (done modify).

Reading Sequences Using HP-IB

An external controller can read the commands in any sequence (in HP-IB command format).

Send the following command to the analyzer:

OUTPSEQx where x is a number from 1 to 6.

Allocate an adequate amount of string variable space in the external controller and execute an

ENTER statement.

6-144 Application and Operation Concepts

The · Special Commands ·

Page 412 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 412 · click to enlarge
Page 412

Ampli er T esting

The HP 8753D allows you to measure the transmission and re ection characteristics of many

ampli ers and active devices . Y ou can measure scalar parameters such as gain, gain atness ,

gain compression, reverse isolation, return loss (SWR), and gain drift versus time . A dditionally ,

you can measure vector parameters such as deviation from linear phase , group delay , complex

impedance and AM-to-PM conversion.

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 412)
Fig. 195 · Chapter 6 Application and Operation Concepts

Figure 6-81. Ampli er P arameters

The analyzer allows you to make a swept-frequency measurement of an ampli er's second or

third harmonic as shown in Figure 6-82.

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 412)
Fig. 196 · Chapter 6 Application and Operation Concepts

Figure 6-82.

Swept Frequency Ampli er Measurement of Absolute Fundamental, 2nd and 3rd

Application and Operation Concepts 6-145

Ampli er parameters · Harmonic Output Levels

Page 413 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 413 · click to enlarge
Page 413

The second/third harmonic response can be displayed directly in dBc , or dB below the

fundamental or carrier (see Figure 6-83). The ability to display harmonic level versus frequency

or RF power allows \real-time" tuning of harmonic distortion. Further , this swept harmonic

measurement, as well as all of the traditional linear ampli er measurements can be made

without reconnecting the test device to a di erent test con guration.

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 413)
Fig. 197 · Chapter 6 Application and Operation Concepts

Figure 6-83.

Swept Frequency Ampli er Measurement of 2nd and 3rd Harmonic Distortion (dBc)

V ector network analyzers are commonly used to characterize ampli er gain compression versus

frequency and power level. This is essentially linear characterization since only the relative

level of the fundamental input to the fundamental output is measured. The narrowband

receiver is tuned to a precise frequency and, as a result, is immune from harmonic distortion.

Y ou may want to quantify the harmonic distortion itself.

Gain compression occurs when the input power of an ampli er is increased to a level that

reduces the gain of the ampli er and causes a nonlinear increase in output power . The point at

which the gain is reduced by 1 dB is called the 1 dB compression point. The gain compression

will vary with frequency , so it is necessary to nd the worst case point of gain compression in

the frequency band.

Once that point is identi ed, you can perform a power sweep of that CW frequency to measure

the input power at which the 1 dB compression occurs and the absolute power out (in dBm) at

compression.

6-146 Application and Operation Concepts

Gain Compression

Page 414 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 414 · click to enlarge
Page 414

Figure 6-84. Diagram of Gain Compression

Figure 6-85 illustrates a simultaneous measurement of fundamental gain compression and

second harmonic power as a function of input power .

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 414)
Fig. 198 · Chapter 6 Application and Operation Concepts

Figure 6-85.

Swept P ower Measurement of Ampli er's Fundamental Gain Compression and 2nd

In a compression measurement it is necessary to know the RF input or output power at a

certain level of gain compression. Therefore , both gain and absolute power level need to be

accurately characterized. Uncertainty in a gain compression measurement is typically less

than 0.05 dB . Also , each input channel of the analyzer is calibrated to display absolute power

(typically within +0.5 dBm up to 3 GHz, and +1 dB up to 6 GHz).

Application and Operation Concepts 6-147

Harmonic Output Level

Page 415 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 415 · click to enlarge
Page 415

Metering the power level

When you are measuring a device that is very sensitive to absolute power level, it is important

that you accurately set the power level at either the device input or output. The analyzer

is capable of using an external HP-IB power meter and controlling source power directly .

Figure 6-86 shows a typical test con guration for setting a precise leveled input power at the

device input.

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 415)
Fig. 199 · Chapter 6 Application and Operation Concepts

Figure 6-86.

T est Con guration for Setting RF Input using A utomatic P ower Meter Calibration

6-148 Application and Operation Concepts

Page 416 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 416 · click to enlarge
Page 416

Mixers or frequency converters , by de nition, exhibit the characteristic of having di erent

input and output frequencies . Mixer tests can be performed using the frequency o set

operation of the analyzer (with an external LO source) or using the tuned receiver operation of

the analyzer (with an external RF and LO source). The most common and convenient method

used is frequency o set.

F or a single-sideband mixer measurement, the RF source can be o set in frequency from the

input receiver frequency , allowing for a swept RF stimulus over one frequency range and

measurement of the IF response over another (in this case the output IF).

T o use the frequency o set guided setup for con guring a mixer measurement:

1. Enter the IF and LO frequencies .
2. Set the LO source to the entered LO frequencies .
3. Specify up conversion or down conversion.
4. Select an RF that is higher or lower in frequency than the LO . (The RF frequencies needed

are calculated by the analyzer .)

The analyzer's tuned receiver mode allows you to tune its receiver to an arbitrary frequency

and measure signal power . This is only possible if the signal you want to analyze is at an exact

known frequency . Therefore , the RF and LO sources must be synthesized and synchronized

with the analyzer time base . Since the analyzer is not phaselocking in this con guration,

you can use it to measure conversion loss of a microwave mixer with an RF frequency range

output.

Y ou must take care to lter the output of the mixer because some of the

intermodulation and leakage products may be very close in frequency to the

desired IF . If these components are not ltered o , the analyzer may have

diculty selecting the correct signal to measure .

Tuned receiver mode also increases dynamic range . Broadband techniques like diode detection

have a high noise oor , while narrowband techniques like tuned receivers are much less

susceptible to noise .

Application and Operation Concepts 6-149

Mixer T esting · Frequency O set · Tuned Receiver

Page 417 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 417 · click to enlarge
Page 417

Mixer P arameters That Y ou Can Measure

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 417)
Fig. 200 · Chapter 6 Application and Operation Concepts

Figure 6-87. Mixer P arameters

Transmission characteristics include conversion loss , conversion compression, group delay ,

and RF feedthru.

Re ection characteristics include return loss , SWR and complex impedance .

Characteristics of the signal at the output port include the output power , the spurious or

harmonic content of the signal, and intermodulation distortion.

Other parameters of concern are isolation terms , including LO to RF isolation and LO to IF

isolation.

In mixer transmission measurements , you have RF and LO inputs and an IF output. Also

emanating from the IF port are several other mixing products of the RF and LO signals . In

mixer re ection measurements , leakage signals from one mixer port propagate and appear

at the other two mixer ports . These unwanted mixing products or leakage signals can cause

distortion by mixing with a harmonic of the analyzer's rst down-conversion stage . T o ensure

that measurement accuracy is not degraded, you must lter certain frequencies or avoid them

by frequency selection. If you place attenuators at all mixer ports , you can reduce mismatch

uncertainties .

6-150 Application and Operation Concepts

A ccuracy Considerations

Page 418 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 418 · click to enlarge
Page 418

Attenuation at Mixer P orts

Mismatch between the instruments , cables , and mixer introduces errors in the measurement

that you cannot remove with a frequency response calibration. Y ou can reduce the mismatch

by using high quality attenuators as close to the mixer under test as possible .

When characterizing linear devices , you can use vector accuracy enhancement (measurement

calibration) to mathematically remove all systematic errors from the measurement, including

source and load mismatches . This is not possible when the device you are characterizing is a

mixer operating over multiple frequency ranges: therefore , source and load mismatches are not

corrected for and will add to overall measurement uncertainty .

T o reduce the measurement errors associated with the interaction between mixer port matches

and system port matches , you can place attenuators at all of the mixer's ports . Figure 6-88

shows a plot of swept conversion loss where no attenuation at mixer ports was used. The

ripple versus frequency is due to source and load mismatches .

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 418)
Fig. 201 · Chapter 6 Application and Operation Concepts

Figure 6-88.

Conversion Loss versus Output Frequency Without Attenuators at

In contrast, Figure 6-90 made use of appropriate attenuation at all mixer ports . Y ou should give

extra care to the selection of the attenuator located at the mixer's IF port to avoid overdriving

the receiver . F or best results , choose the value of this attenuator so that the power incident on

the analyzer's R channel port is less than 10 dBm and greater than 35 dBm.

Application and Operation Concepts 6-151

Mixer P orts

Page 419 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 419 · click to enlarge
Page 419

Harmonics , linearity , and spurious signals also introduce errors that are not removed by

frequency response calibration. These errors are smaller with a narrowband detection scheme ,

but they may still interfere with your measurements . Y ou should lter the IF signal to reduce

these errors as much as possible .

Correct ltering between the mixer's IF port and the receiver's input port can eliminate

unwanted mixing and leakage signals from entering the analyzer's receiver . Figure 6-89 shows

a plot of mixer conversion loss when proper IF ltering was neglected.

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 419)
Fig. 202 · Chapter 6 Application and Operation Concepts

Figure 6-89.

Example of Conversion Loss versus Output Frequency Without Correct

IF Signal P ath Filtering

Figure 6-90 shows the same mixer's conversion loss with the addition of a low pass lter at the

mixer's IF port.

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 419)
Fig. 203 · Chapter 6 Application and Operation Concepts

Figure 6-90.

Example of Conversion Loss versus Output Frequency With Correct IF

Signal P ath Filtering and Attenuation at all Mixer P orts

Filtering is required in both xed and broadband measurements , but you can implement it more

easily in the xed situation. Therefore , when con guring broad-band (swept) measurements ,

you may need to trade some measurement bandwidth for the ability to more selectively lter

signals entering the analyzer's receiver .

6-152 Application and Operation Concepts

Filtering

Page 420 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNN
Page 420 · click to enlarge
Page 420

By choosing test frequencies (frequency list mode), you can reduce the e ect of spurious

responses on measurements by avoiding frequencies that produce IF signal path distortion.

LO Frequency A ccuracy and Stability

The analyzer source is phaselocked to its receiver through a reference loop . In the frequency

o set mode , the mixer under test is inserted in this loop . T o ensure that the analyzer

phaselocks correctly , it is important that you use an LO source that has frequency accuracy

better than 1 MHz and residual FM < 20 kHz RMS.

Up-Conversion and Down-Conversion De nition

NNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNN

When you choose between

RF < LO and

RF > LO in the frequency o set menus , the analyzer

determines which direction the internal source must sweep in order to achieve the requested

IF frequency . F or example , to measure the lower sideband of a mixer , where the RF signal

is below the LO (RF < LO), the internal source must sweep backwards . See the examples in

Figure 6-91.

Agilent Technologies 8753D — NNNNNNNNNNNNNNNNNNNNNNN diagram (page 420)
Fig. 204

Figure 6-91. Examples of Up Converters and Down Converters

In standard mixer measurements , the input of the mixer is always connected to the analyzer's

RF source , and the output of the mixer always produces the IF frequencies that are received by

the analyzer's receiver .

Application and Operation Concepts 6-153

Frequency Selection

Page 421 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNN
Page 421 · click to enlarge
Page 421

However , the ports labeled RF and IF on most mixers are not consistently connected to the

analyzer's source and receiver ports , respectively . These mixer ports are switched, depending

on whether a down converter or an upconverter measurement is being performed.

It is important to keep in mind that in the setup diagrams of the frequency o set mode , the

analyzer's source and receiver ports are labeled according to the mixer port that they are

connected to .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

In a down converter measurement where the

DOWN CONVERTER softkey is selected, the

labeled RF , connecting to the mixer RF port, and the analyzer's receiver frequency is labeled

IF , connecting to the mixer IF port.

Because the RF frequency can be greater or less than the set LO frequency in this type of

NNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNN

measurement, you can select either

RF > LO or

RF < LO .

Figure 6-92. Down Converter P ort Connections

6-154 Application and Operation Concepts

Page 422 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNN
Page 422 · click to enlarge
Page 422

In an up converter measurement where the

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

UP CONVERTER softkey is selected, the notation

on the setup diagram indicates that the analyzer's source frequency is labeled IF , connecting

to the mixer IF port, and the analyzer's receiver frequency is labeled RF , connecting to the

mixer RF port.

Because the RF frequency will always be greater than the set LO frequency in this type of

NNNNNNNNNNNNNNNNNNNNNNN

measurement, you must select only

RF > LO .

Figure 6-93. Up Converter P ort Connections

Application and Operation Concepts 6-155

Page 423 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 423 · click to enlarge
Page 423

Figure 6-94.

Example Spectrum of RF , LO , and IF Signals Present in a Conversion Loss Measurement

Conversion loss is a measure of how eciently a mixer converts energy from one frequency

to another . It is the ratio of the sideband output power to input signal power and is usually

expressed in dB .

Since the frequency response of the test system gets measured with the mixer's response , you

can perform a frequency response calibration to remove this group of errors .

Figure 6-95. Main Isolation T erms

Isolation is the amount of attenuation provided when a signal is applied to one port of a mixer

and measured at another port. Figure 6-95 shows the three main isolation terms .

LO F eedthru / LO to RF Leakage

LO feedthru, or LO-to-IF isolation, is the amount the LO signal that is attenuated when it

reaches the IF port.

LO to RF isolation is the amount the LO power is attenuated when it appears directly at the RF

port.

Both of these LO isolation terms are small for single and double balanced mixers . The RF signal

level applied to the mixer will have an a ect on this measurement. F or this reason, these

terms are usually measured with the RF port of the mixer terminated in a matched state .

6-156 Application and Operation Concepts

Conversion Loss · Isolation

Page 424 of the Agilent Technologies 8753D user manual — S W R =
Page 424 · click to enlarge
Page 424

RF feedthru, or RF-to-IF isolation, is the amount the RF power that is attenuated when it

reaches the IF port. This value is low in double balanced mixers . RF feedthru is usually less of

a problem than the LO isolation terms because the LO power level is signi cantly higher than

the RF power drive .

Y ou can make an RF feedthru measurement using the same instruments and setup that you use

to measure conversion loss . Because the source and receiver frequencies are the same , the

analyzer can use narrowband (tuned receiver) detection to make the measurement. The only

di erence that you need in the hardware con guration is that the IF lter needs to be removed

so the RF feedthru will not be ltered out.

The RF signal is applied to the RF port of the mixer and the feedthru is measured at the IF

port.

The RF feedthru level is very dependent on the LO signal that is applied. F or this reason, you

should make the measurement with the LO signal present at its normal operating level.

Y ou should perform a frequency response calibration to improve accuracy .

SWR / Return Loss

Re ection coecient () is de ned as the ratio between the re ected voltage (Vr) and incident

voltage (Vi). Standing wave ratio (SWR) is de ned as the ratio of maximum standing wave

voltage to the minimum standing wave voltage and can be derived from the re ection

coecient () using the equation shown below . Return loss can be derived from the re ection

coecient as well.

1 + j j

S W R =

1 j j

R etur n l oss = 20 l og j j

Mixers are three-port devices , and the re ection from any one port

depends on the conditions of the other two ports . Y ou should replicate the

operating conditions the mixer will experience as closely as possible for the

measurement.

When you measure the RF port SWR, you should have the LO drive level present and set to the

expected frequency and power levels . The IF port should be terminated in a condition as close

to its operating state as possible .

The measurements of LO port SWR and IF port SWR are very similar . F or IF port SWR, you

should terminate the RF port in a matched condition and apply the LO signal at its normal

operating level. F or the LO port SWR, the RF and IF ports should both be terminated in

conditions similar to what will be present during normal operation.

Application and Operation Concepts 6-157

RF F eedthru · = · V r

Page 425 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 425 · click to enlarge
Page 425

Figure 6-96. Conversion Loss and Output P ower as a Function of Input P ower Level

Conversion compression is a measure of the maximum RF input signal level for which the

mixer will provide linear operation. The conversion loss is the ratio of the IF output level to

the RF input level, and this value remains constant over a speci ed input power range . When

the input power level exceeds a certain maximum, the constant ratio between IF and RF power

levels will begin to change . The point at which the ratio has decreased 1 dB is called the 1 dB

compression point.

level, until mixer compression begins and the mixer saturates .

Y ou can measure conversion compression using the same basic test con gurations that are used

to measure the conversion loss .

T o set up for a conversion compression measurement, rst measure the conversion loss of the

mixer under test. Set up for a CW measurement at the frequency of interest. Sweeping the

RF drive level over a 25 dB span soon shows the power level at which the conversion loss

increases by 1 dB .

With power meter calibration controlling the RF drive level, and the receiver calibrated to

measure output power , you can make accurate measurements of the output power at the 1 dB

compression point.

When you are making linear measurements , provide a reference for determining phase by

splitting the RF source power and send part of the signal into the reference channel. (This does

not work for frequency o set measurements , since the source and receiver are functioning at

di erent frequencies .)

T o provide a reference signal for the phase measurement, you need a second mixer . This mixer

is driven by the same RF and LO signals that are used to drive the mixer under test. The IF

output from the reference mixer is applied to the reference (R) channel of the analyzer .

6-158 Application and Operation Concepts

Conversion Compression · Phase Measurements

Page 426 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 426 · click to enlarge
Page 426

Amplitude and Phase Tracking

The match between mixers is de ned as the absolute di erence in amplitude and/or phase

response over a speci ed frequency range . The tracking between mixers is essentially how well

the devices are matched over a speci ed interval. This interval may be a frequency interval or

a temperature interval, or a combination of both.

Y ou can make tracking measurements by ratioing the responses of two mixer conversion loss

measurements . Then any di erence you view in response is due to the mixers and not the

measurement system.

Replace mixer A with the mixer that you want to compare it to . Mixer R should always remain

in place as the reference mixer .

Figure 6-97.

Connections for an Amplitude and Phase Tracking Measurement Between Two Mixers

Phase Linearity and Group Delay

Group delay is the rate of change of phase through a device with respect to frequency (d /d! ).

Traditionally , group delay has been used to describe the propagation delay ( g), and deviation

from linear phase through a linear device . However , this parameter also contains valuable

information about transmission delay and distortion through a non-linear device such as a

mixer or frequency converter . F or example , at group delay corresponds to low modulation

distortion (that is , carrier and sidebands propagate at the same rate).

Phase linearity and group delay are both measurements of the distortion of a transmitted

signal. Both measure the non-linearity of a device's phase response with respect to frequency .

Application and Operation Concepts 6-159

Page 427 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 427 · click to enlarge
Page 427

In standard vector error-correction, a thru (delay=0) is used as a calibration standard. The

solution to this problem is to use a calibration mixer with very small group delay as the

calibration standard.

An important characteristic to remember when selecting a calibration mixer is that the delay

of the device should be kept as low as possible . T o do this , select a mixer with very wide

bandwidth (wider bandwidth results in smaller delay).

6-160 Application and Operation Concepts

Page 428 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 428 · click to enlarge
Page 428

T o minimize the error introduced when you add an adapter to a measurement system, the

adapter needs to have low SWR or mismatch, low loss , and high repeatability .

Agilent Technologies 8753D — Chapter 6 Application and Operation Concepts diagram (page 428)
Fig. 205 · Chapter 6 Application and Operation Concepts

Figure 6-98. A dapter Considerations

In a re ection measurement, the directivity of a system is a measure of the error introduced

by an imperfect signal separation device . It typically includes any signal that is detected at the

coupled port which has not been re ected by the test device . This directivity error will add

with the true re ected signal from the device , causing an error in the measured data. Overall

directivity is the limit to which a device's return loss or re ection can be measured. Therefore ,

it is important to have good directivity to measure low re ection devices .

F or example , a coupler has a 7 mm connector and 40 dB directivity , which is equivalent to a

re ection coecient of =0.01 (directivity in dB = 20 log ). Suppose we want to connect to

a device with an SMA male connector . W e need to adapt from 7 mm to SMA.

If we choose a precision 7 mm to SMA adapter with a SWR of 1.06, which has =0.03, the

overall directivity becomes =0.04 or 28 dB . However , if we use two adapters to do the same

job , the re ection from each adapter adds up to degrade the directivity to 17 dB . The last

example shown in Figure 6-98 uses three adapters that shows an even worse directivity of

14 dB . It is clear that a low SWR is desirable to avoid degrading the directivity of the system.

Application and Operation Concepts 6-161

Connection Considerations · A dapters

Page 429 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 429 · click to enlarge
Page 429

Fixtures are needed to interface non-coaxial devices to coaxial test instruments . It may also be

necessary to transform the characteristic impedance from standard 50 ohm instruments to a

non-standard impedance and to apply bias if an active device is being measured.

F or accurate measurements , the xture must introduce minimum change to the test signal, not

destroy the test device , and provide a repeatable connection to the device .

Hewlett-P ackard o ers several xtures for TO cans , stripline , and microstrip devices . Refer to

Chapter 11, \Compatible P eripherals ."

If Y ou W ant to Design Y our Own Fixture

Ideally , a xture should provide a transparent connection between the test instrument and

the test device . This means it should have no loss or electrical length and a at frequency

response , to prevent distortion of the actual signal. A perfect match to both the instrument

and the test device eliminates re ected test signals . The signal should be e ectively coupled

into the test device , rather than leaking around the device and resulting in crosstalk from input

to output. Repeatable connections are necessary to ensure consistent data.

Realistically , it is impossible to build an ideal xture , especially at high frequencies . However ,

it is possible to optimize the performance of the test xture relative to the performance of

the test device . If the xture's e ects on the test signal are relatively small compared to the

device's parameters , then the xture's e ects can be assumed to be negligible .

F or example , if the xture's loss is much less than the acceptable measurement uncertainty at

the test frequency , then it can be ignored.

6-162 Application and Operation Concepts

Fixtures

Page 430 of the Agilent Technologies 8753D user manual — 5091-8804E
Page 430 · click to enlarge
Page 430

Hewlett-P ackard Company , \Simplify Y our Ampli er and Mixer T esting" 5956-4363

Hewlett-P ackard Company , \RF and Microwave Device T est for the '90s Seminar P apers"

5091-8804E

Hewlett-P ackard Company \T esting Ampli ers and A ctive Devices with the HP 8720 Network

Analyzer" Product Note 8720-1 5091-1942E

Hewlett-P ackard Company \Mixer Measurements Using the HP 8753 Network Analyzer"

Product Note 8753-2A 5952-2771

General Measurement and Calibration T echniques

Rytting, Doug, \E ects of Uncorrected RF P erformance in a V ector Network Analyzer ," from

\Microwave Journal," April 1991

Blacka, Robert J ., \TDR Gated Measurements of Stripline T erminations ," Reprint from

\Microwave Product Digest," HP publication no . 5952-0359, March/April 1989

Montgomery , David, \Borrowing RF T echniques for Digital Design," Reprint from \Computer

Design" HP publication number 5952-9335, May 1982

Rytting, Doug, \A dvances in Microwave Error Correction T echniques ," Hewlett-P ackard RF and

Microwave Measurement Symposium paper HP publication number 5954-8378, June 1987

Rytting, Doug, \Improved RF Hardware and Calibration Methods for Network Analyzers ,"

Hewlett-P ackard RF and Microwave Measurement Symposium paper , 1991

Dunsmore , Joel, \A dd P ower-Meter A ccuracy to a Network Analyzer ," from \Microwaves and

RF ," January 1991

Fixtures and Non-Coaxial Measurements

Hewlett-P ackard Company , \Applying the HP 8510 TRL Calibration for Non-Coaxial

Measurements ," Product Note 8510-8A HP publication number 5091-3645E, F ebruary 1992

Hewlett-P ackard Company , \Measuring Chip Capacitors with the HP 8520C Network Analyzers

and Inter-Continental Microwave T est Fixtures" Product Note 8510-17 HP publication number

Hewlett-P ackard Company , \In-Fixture Microstrip Device Measurement Using TRL*

Calibration," Product Note 8720-2 HP publication number 5091-1943E, A ugust 1991

Hewlett-P ackard Company , \Calibration and Modeling Using the HP 83040 Modular Microcircuit

P ackage ," Product Note 83040-2 HP publication number 5952-1907, May 1990

\T est Fixtures and Calibration Standards ," Inter-Continental Microwave Product Catalog HP

Curran, Jim, \Network Analysis of Fixtured Devices ," Hewlett-P ackard RF and Microwave

Measurement Symposium paper , HP publication number 5954-8346, September 1986

Curran, Jim, \TRL Calibration for Non-Coaxial Measurements ," Hewlett-P ackard Semiconductor

T est Symposium paper

Elmore , Glenn and Louis Salz, \Quality Microwave Measurement of P ackaged A ctive Devices ,"

Hewlett-P ackard Journal, F ebruary 1987

\Measurement T echniques for Fixtured Devices ," HP 8510/8720 News HP publication number

Application and Operation Concepts 6-163

Show 4 more labels / short text on this page
Reference Documents · 5091-5674E, September 1992 · publication number 5091-4254E · 5952-2766, June 1990
Page 431 of the Agilent Technologies 8753D user manual — Chapter 6 Application and Operation Concepts
Page 431 · click to enlarge
Page 431

Hewlett-P ackard Company , \On-W afer Measurements Using the HP 8510 Network Analyzer and

Cascade Microtech W afer Probes ," Product Note 8510-6 HP publication number 5954-1579

Barr , J .T ., T . Burcham, A.C. Davidson, E.W . Strid, \A dvancements in On-W afer Probing

Calibration T echniques ," Hewlett-P ackard RF and Microwave Measurement Symposium paper ,

Lautzenhiser , S., A. Davidson, D . Jones , \Improve A ccuracy of On-W afer T ests Via LRM

Calibration," Reprinting from \Microwaves and RF" HP publication number 5952-1286, January

\On-W afer CalibrationPractical Considerations ," HP 8510/8720 News HP publication number

5091-6837, F ebruary 1993

6-164 Application and Operation Concepts

On-W afer Measurements

Page 432 of the Agilent Technologies 8753D user manual — Chapter 7 Specifications and Measurement Uncertainties
Page 432 · click to enlarge
Page 432

Speci cations and Measurement Uncertainties

The speci cations described in the table below apply to transmission measurements using 10 Hz

IF BW and full 2-port correction. Dynamic range is limited by the maximum test port power

and the receiver's noise oor .

T able 7-1. HP 8753D Dynamic Range

Frequency Range Dynamic Range

30 kHz to 300 kHz 100 dB

300 kHz to 1.3 GHz 110 dB

1.3 GHz to 3 GHz 110 dB

3 GHz to 6 GHz 105 dB

* 90 dB , 30 kHz to 50 kHz

y 100 dB , 300 kHz to 16 MHz, due to xed spurs

z 105 dB , Option 075

Speci cations and Measurement Uncertainties 7-1

Show 4 more labels / short text on this page
Dynamic Range · * · y z · z
Page 433 of the Agilent Technologies 8753D user manual — Chapter 7 Specifications and Measurement Uncertainties
Page 433 · click to enlarge
Page 433

HP 8753D Network Analyzer Speci cations

HP 8753D (50 ) with 7 mm T est P orts

The following speci cations describe the system performance of the HP 8753D network

analyzer . The system hardware includes the following:

Options: 006

Calibration kit: HP 85031B

Cables: HP 11857D

Measurement P ort Characteristics

The following tables describe the measurement port characteristics for both corrected and

uncorrected HP 8753D network analyzers .

T able 7-2.

Measurement P ort Characteristics (Corrected

) for HP 8753D (50 ) with

7 mm T est P orts

30 kHz to 300 kHz

300 kHz to 1.3 GHz 1.3 GHz to 3 GHz 3 GHz to 6 GHz

Directivity 55 dB 55 dB 51 dB 46 dB

Source match 55 dB 51 dB 49 dB 43 dB

Load match 55 dB 55 dB 51 dB 46 dB

Re ection tracking 0.001 dB 0.001 dB 0.005 dB 0.020 dB

Transmission tracking 0.008 dB 0.006 dB 0.009 dB 0.021 dB

* These characteristics apply for an environmental temperature of 25 5

C, with less than 1

the calibration temperature .

T able 7-3.

Measurement P ort Characteristics (Uncorrected

) for HP 8753D (50 )

with 7 mm T est P orts

30 kHz to 300 kHz

300 kHz to 1.3 GHz 1.3 GHz to 3 GHz 3 GHz to 6 GHz

35 dB 30 dB 25 dB

Source match 18 dB

16 dB 16 dB 14 dB

Load match 20 dB

18 dB 16 dB 14 dB

Re ection tracking 2.0 dB 1.5 dB 1.5 dB 2.5 dB

Transmission tracking 2.0 dB 1.5 dB 1.5 dB 2.5 dB

Crosstalk 100 dB 100 dB 100 dB 90 dB

Applies at 25 5

10 dB , 30 kHz to 50 kHz

15 dB , 30 kHz to 50 kHz

7-2 Speci cations and Measurement Uncertainties

Show 18 more labels / short text on this page
* · Frequency Range · y · · · C deviation from · y · Typical P erformance · * · Frequency Range · y · z · Directivity 20 dB · * · · y · Typical · z
Page 434 of the Agilent Technologies 8753D user manual — Chapter 7 Specifications and Measurement Uncertainties
Page 434 · click to enlarge
Page 434
Agilent Technologies 8753D — Chapter 7 Specifications and Measurement Uncertainties diagram (page 434)
Fig. 206 · Chapter 7 Specifications and Measurement Uncertainti…
Agilent Technologies 8753D — Chapter 7 Specifications and Measurement Uncertainties diagram (page 434)
Fig. 207 · Chapter 7 Specifications and Measurement Uncertainti…

Speci cations and Measurement Uncertainties 7-3

Transmission Measurement Uncertainties

Page 435 of the Agilent Technologies 8753D user manual — Chapter 7 Specifications and Measurement Uncertainties
Page 435 · click to enlarge
Page 435

Re ection Measurement Uncertainties

Agilent Technologies 8753D — Chapter 7 Specifications and Measurement Uncertainties diagram (page 435)
Fig. 208 · Chapter 7 Specifications and Measurement Uncertainti…
Agilent Technologies 8753D — Chapter 7 Specifications and Measurement Uncertainties diagram (page 435)
Fig. 209 · Chapter 7 Specifications and Measurement Uncertainti…

7-4 Speci cations and Measurement Uncertainties

Page 436 of the Agilent Technologies 8753D user manual — Chapter 7 Specifications and Measurement Uncertainties
Page 436 · click to enlarge
Page 436

HP 8753D (50 ) with Type-N T est P orts

The following speci cations describe the system performance of the HP 8753D network

analyzer . The system hardware includes the following:

Options: 006

Calibration kit: HP 85032B/E

Cables: HP 11851B

Measurement P ort Characteristics

The following table describes the measurement port characteristics for corrected HP 8753D

network analyzers .

T able 7-4.

Measurement P ort Characteristics (Corrected)

for HP 8753D (50 ) with

Type-N T est P orts

30 kHz to 300 kHz

300 kHz to 1.3 GHz 1.3 GHz to 3 GHz 3 GHz to 6 GHz

Directivity 50 dB 50 dB 47 dB 40 dB

Source match 49 dB 42 dB 36 dB 31 dB

Load match 50 dB 50 dB 47 dB 40 dB

Re ection tracking 0.005 dB 0.009 dB 0.019 dB 0.070 dB

Transmission tracking 0.014 dB 0.013 dB 0.026 dB 0.065 dB

Applies at 25 5

Speci cations and Measurement Uncertainties 7-5

Show 7 more labels / short text on this page
* · Frequency Range · y · * · · y · Typical performance
Page 437 of the Agilent Technologies 8753D user manual — Chapter 7 Specifications and Measurement Uncertainties
Page 437 · click to enlarge
Page 437
Agilent Technologies 8753D — Chapter 7 Specifications and Measurement Uncertainties diagram (page 437)
Fig. 210 · Chapter 7 Specifications and Measurement Uncertainti…
Agilent Technologies 8753D — Chapter 7 Specifications and Measurement Uncertainties diagram (page 437)
Fig. 211 · Chapter 7 Specifications and Measurement Uncertainti…

7-6 Speci cations and Measurement Uncertainties

Transmission Measurement Uncertainties

Page 438 of the Agilent Technologies 8753D user manual — Chapter 7 Specifications and Measurement Uncertainties
Page 438 · click to enlarge
Page 438

Re ection Measurement Uncertainties

Agilent Technologies 8753D — Chapter 7 Specifications and Measurement Uncertainties diagram (page 438)
Fig. 212 · Chapter 7 Specifications and Measurement Uncertainti…
Agilent Technologies 8753D — Chapter 7 Specifications and Measurement Uncertainties diagram (page 438)
Fig. 213 · Chapter 7 Specifications and Measurement Uncertainti…

Speci cations and Measurement Uncertainties 7-7

Page 439 of the Agilent Technologies 8753D user manual — Chapter 7 Specifications and Measurement Uncertainties
Page 439 · click to enlarge
Page 439

HP 8753D (50 ) with 3.5 mm T est P orts

The following speci cations describe the system performance of the HP 8753D network

analyzer . The system hardware includes the following:

Options: 006

Calibration kit: HP 85033D

Cables: HP 11857D

Measurement P ort Characteristics

The following table describes the measurement port characteristics for corrected HP 8753D

network analyzers .

T able 7-5.

Measurement P ort Characteristics (Corrected)

for HP 8753D (50 ) with

3.5 mm T est P orts

30 kHz to 300 kHz

300 kHz to 1.3 GHz 1.3 GHz to 3 GHz 3 GHz to 6 GHz

Directivity 49 dB 46 dB 44 dB 38 dB

Source match 49 dB 44 dB 41 dB 37 dB

Load match 49 dB 46 dB 44 dB 38 dB

Re ection tracking 0.010 dB 0.005 dB 0.007 dB 0.009 dB

Transmission tracking 0.016 dB 0.014 dB 0.022 dB 0.048 dB

Applies at 25 5

7-8 Speci cations and Measurement Uncertainties

Show 7 more labels / short text on this page
* · Frequency Range · y · * · · y · Typical P erformance
Page 440 of the Agilent Technologies 8753D user manual — Chapter 7 Specifications and Measurement Uncertainties
Page 440 · click to enlarge
Page 440
Agilent Technologies 8753D — Chapter 7 Specifications and Measurement Uncertainties diagram (page 440)
Fig. 214 · Chapter 7 Specifications and Measurement Uncertainti…
Agilent Technologies 8753D — Chapter 7 Specifications and Measurement Uncertainties diagram (page 440)
Fig. 215 · Chapter 7 Specifications and Measurement Uncertainti…

Speci cations and Measurement Uncertainties 7-9

Transmission Measurement Uncertainties

Page 441 of the Agilent Technologies 8753D user manual — Chapter 7 Specifications and Measurement Uncertainties
Page 441 · click to enlarge
Page 441

Re ection Measurement Uncertainties

Agilent Technologies 8753D — Chapter 7 Specifications and Measurement Uncertainties diagram (page 441)
Fig. 216 · Chapter 7 Specifications and Measurement Uncertainti…
Agilent Technologies 8753D — Chapter 7 Specifications and Measurement Uncertainties diagram (page 441)
Fig. 217 · Chapter 7 Specifications and Measurement Uncertainti…

7-10 Speci cations and Measurement Uncertainties

Page 442 of the Agilent Technologies 8753D user manual — Chapter 7 Specifications and Measurement Uncertainties
Page 442 · click to enlarge
Page 442

HP 8753D (75 ) with Type-N T est P orts

The following speci cations describe the system performance of the HP 8753D network

analyzer . The system hardware includes the following:

Options: 075

Calibration kit: HP 85036B

Cables: HP 11857B

Measurement P ort Characteristics

The following tables describe the measurement port characteristics for both corrected and

uncorrected HP 8753D network analyzers .

T able 7-6.

Measurement P ort Characteristics (Corrected)

for HP 8753D (75 ) with

Type-N T est P orts

30 kHz to 300 kHz

300 kHz to 1.3 GHz 1.3 GHz to 3 GHz

Directivity 48 dB 48 dB 43 dB

Source match 47 dB 41 dB 35 dB

Load match 48 dB 48 dB 43 dB

Re ection tracking 0.004 dB 0.010 dB 0.019 dB

Transmission tracking 0.018 dB 0.015 dB 0.033 dB

Applies at 25 5

T able 7-7.

Measurement P ort Characteristics (Uncorrected)

for HP 8753D (75 )

with Type-N T est P orts

30 kHz to 300 kHz

300 kHz to 1.3 GHz 1.3 GHz to 3 GHz

Directivity 34 dB 35 dB 30 dB

Source match 10 dB 16 dB 16 dB

Load match 14 dB 18 dB 16 dB

Re ection tracking 2.0 dB 1.5 dB 1.5 dB

Transmission tracking 2.0 dB 1.5 dB 1.5 dB

Crosstalk 100 dB 100 dB 100 dB

Applies at 25 5

Speci cations and Measurement Uncertainties 7-11

Show 14 more labels / short text on this page
* · Frequency Range · y · * · · y · Typical P erformance · * y · Frequency Range · y · * · · y · Typical P erformance
Page 443 of the Agilent Technologies 8753D user manual — Chapter 7 Specifications and Measurement Uncertainties
Page 443 · click to enlarge
Page 443
Agilent Technologies 8753D — Chapter 7 Specifications and Measurement Uncertainties diagram (page 443)
Fig. 218 · Chapter 7 Specifications and Measurement Uncertainti…
Agilent Technologies 8753D — Chapter 7 Specifications and Measurement Uncertainties diagram (page 443)
Fig. 219 · Chapter 7 Specifications and Measurement Uncertainti…

7-12 Speci cations and Measurement Uncertainties

Transmission Measurement Uncertainties

Page 444 of the Agilent Technologies 8753D user manual — Chapter 7 Specifications and Measurement Uncertainties
Page 444 · click to enlarge
Page 444

Re ection Measurement Uncertainties

Agilent Technologies 8753D — Chapter 7 Specifications and Measurement Uncertainties diagram (page 444)
Fig. 220 · Chapter 7 Specifications and Measurement Uncertainti…
Agilent Technologies 8753D — Chapter 7 Specifications and Measurement Uncertainties diagram (page 444)
Fig. 221 · Chapter 7 Specifications and Measurement Uncertainti…

Speci cations and Measurement Uncertainties 7-13

Page 445 of the Agilent Technologies 8753D user manual — Chapter 7 Specifications and Measurement Uncertainties
Page 445 · click to enlarge
Page 445

HP 8753D (75 ) with Type-F T est P orts

The following speci cations describe the system performance of the HP 8753D network

analyzer . The system hardware includes the following:

Options: 075

Calibration kit: HP 85039A

Cables: HP 11857B

Measurement P ort Characteristics

The following table describes the measurement port characteristics for corrected HP 8753D

network analyzers .

T able 7-8.

Measurement P ort Characteristics (Corrected)

for HP 8753D (75 )

using HP 85039A F-M T est P orts

30 kHz to 300 kHz

300 kHz to 1.3 GHz 1.3 GHz to 3 GHz

Directivity 38 dB 38 dB 32 dB

Source match 36 dB 36 dB 30 dB

Load match 38 dB 38 dB 32 dB

Re ection tracking 0.0080 dB 0.0080 dB 0.0320 dB

Transmission tracking 0.0618 dB 0.0346 dB 0.0778 dB

Applies at 25 5

7-14 Speci cations and Measurement Uncertainties

Show 7 more labels / short text on this page
* · Frequency Range · y · * · · y · Typical P erformance
Page 446 of the Agilent Technologies 8753D user manual — Chapter 7 Specifications and Measurement Uncertainties
Page 446 · click to enlarge
Page 446
Agilent Technologies 8753D — Chapter 7 Specifications and Measurement Uncertainties diagram (page 446)
Fig. 222 · Chapter 7 Specifications and Measurement Uncertainti…
Agilent Technologies 8753D — Chapter 7 Specifications and Measurement Uncertainties diagram (page 446)
Fig. 223 · Chapter 7 Specifications and Measurement Uncertainti…

Speci cations and Measurement Uncertainties 7-15

Transmission Measurement Uncertainties

Page 447 of the Agilent Technologies 8753D user manual — Chapter 7 Specifications and Measurement Uncertainties
Page 447 · click to enlarge
Page 447

Re ection Measurement Uncertainties

Agilent Technologies 8753D — Chapter 7 Specifications and Measurement Uncertainties diagram (page 447)
Fig. 224 · Chapter 7 Specifications and Measurement Uncertainti…
Agilent Technologies 8753D — Chapter 7 Specifications and Measurement Uncertainties diagram (page 447)
Fig. 225 · Chapter 7 Specifications and Measurement Uncertainti…

7-16 Speci cations and Measurement Uncertainties

Page 448 of the Agilent Technologies 8753D user manual — Chapter 7 Specifications and Measurement Uncertainties
Page 448 · click to enlarge
Page 448

T able 7-9.

Measurement P ort Characteristics (Corrected)

for HP 8753D (75 )

using HP 85039A F-F T estports

30 kHz to 300 kHz

300 kHz to 1.3 GHz 1.3 GHz to 3 GHz

Directivity 38 dB 38 dB 32 dB

Source match 36 dB 36 dB 30 dB

Load match 32 dB 32 dB 26 dB

Re ection tracking 0.0080 dB 0.0080 dB 0.0320 dB

Transmission tracking 0.0959 dB 0.0518 dB 0.1118 dB

Applies at 25 5

Speci cations and Measurement Uncertainties 7-17

Show 7 more labels / short text on this page
* · Frequency Range · y · * · · y · Typical P erformance
Page 449 of the Agilent Technologies 8753D user manual — Chapter 7 Specifications and Measurement Uncertainties
Page 449 · click to enlarge
Page 449
Agilent Technologies 8753D — Chapter 7 Specifications and Measurement Uncertainties diagram (page 449)
Fig. 226 · Chapter 7 Specifications and Measurement Uncertainti…
Agilent Technologies 8753D — Chapter 7 Specifications and Measurement Uncertainties diagram (page 449)
Fig. 227 · Chapter 7 Specifications and Measurement Uncertainti…

7-18 Speci cations and Measurement Uncertainties

Transmission Measurement Uncertainties

Page 450 of the Agilent Technologies 8753D user manual — Chapter 7 Specifications and Measurement Uncertainties
Page 450 · click to enlarge
Page 450

Re ection Measurement Uncertainties

Agilent Technologies 8753D — Chapter 7 Specifications and Measurement Uncertainties diagram (page 450)
Fig. 228 · Chapter 7 Specifications and Measurement Uncertainti…
Agilent Technologies 8753D — Chapter 7 Specifications and Measurement Uncertainties diagram (page 450)
Fig. 229 · Chapter 7 Specifications and Measurement Uncertainti…

Speci cations and Measurement Uncertainties 7-19

Page 451 of the Agilent Technologies 8753D user manual — Chapter 7 Specifications and Measurement Uncertainties
Page 451 · click to enlarge
Page 451

T able 7-10.

Measurement P ort Characteristics (Corrected)

for HP 8753D (75 )

using HP 85039A M-M T estports

30 kHz to 300 kHz

300 kHz to 1.3 GHz 1.3 GHz to 3 GHz

Directivity 32 dB 32 dB 26 dB

Source match 31.950 dB 31.170 dB 25.080 dB

Load match 38 dB 38 dB 32 dB

Re ection tracking 0.0348 dB 0.0253 dB 0.0728 dB

Transmission tracking 0.0780 dB 0.0474 dB 0.1106 dB

Applies at 25 5

7-20 Speci cations and Measurement Uncertainties

Show 7 more labels / short text on this page
* · Frequency Range · y · * · · y · Typical P erformance
Page 452 of the Agilent Technologies 8753D user manual — Chapter 7 Specifications and Measurement Uncertainties
Page 452 · click to enlarge
Page 452
Agilent Technologies 8753D — Chapter 7 Specifications and Measurement Uncertainties diagram (page 452)
Fig. 230 · Chapter 7 Specifications and Measurement Uncertainti…
Agilent Technologies 8753D — Chapter 7 Specifications and Measurement Uncertainties diagram (page 452)
Fig. 231 · Chapter 7 Specifications and Measurement Uncertainti…

Speci cations and Measurement Uncertainties 7-21

Transmission Measurement Uncertainties

Page 453 of the Agilent Technologies 8753D user manual — Chapter 7 Specifications and Measurement Uncertainties
Page 453 · click to enlarge
Page 453

Re ection Measurement Uncertainties

Agilent Technologies 8753D — Chapter 7 Specifications and Measurement Uncertainties diagram (page 453)
Fig. 232 · Chapter 7 Specifications and Measurement Uncertainti…
Agilent Technologies 8753D — Chapter 7 Specifications and Measurement Uncertainties diagram (page 453)
Fig. 233 · Chapter 7 Specifications and Measurement Uncertainti…

7-22 Speci cations and Measurement Uncertainties

Page 454 of the Agilent Technologies 8753D user manual — Chapter 7 Specifications and Measurement Uncertainties
Page 454 · click to enlarge
Page 454

T able 7-11.

Measurement P ort Characteristics (Corrected)

for HP 8753D (75 )

using HP 85039A M-F T estports

30 kHz to 300 kHz

300 kHz to 1.3 GHz 1.3 GHz to 3 GHz

Directivity 32 dB 32 dB 26 dB

Source match 31.950 dB 31.170 dB 25.080 dB

Load match 32 dB 32 dB 26 dB

Re ection tracking 0.0348 dB 0.0253 dB 0.0728 dB

Transmission tracking 0.1121 dB 0.0646 dB 0.1445 dB

Applies at 25 5

Speci cations and Measurement Uncertainties 7-23

Show 7 more labels / short text on this page
* · Frequency Range · y · * · · y · Typical P erformance
Page 455 of the Agilent Technologies 8753D user manual — Chapter 7 Specifications and Measurement Uncertainties
Page 455 · click to enlarge
Page 455
Agilent Technologies 8753D — Chapter 7 Specifications and Measurement Uncertainties diagram (page 455)
Fig. 234 · Chapter 7 Specifications and Measurement Uncertainti…
Agilent Technologies 8753D — Chapter 7 Specifications and Measurement Uncertainties diagram (page 455)
Fig. 235 · Chapter 7 Specifications and Measurement Uncertainti…

7-24 Speci cations and Measurement Uncertainties

Transmission Measurement Uncertainties

Page 456 of the Agilent Technologies 8753D user manual — Chapter 7 Specifications and Measurement Uncertainties
Page 456 · click to enlarge
Page 456

Re ection Measurement Uncertainties

Agilent Technologies 8753D — Chapter 7 Specifications and Measurement Uncertainties diagram (page 456)
Fig. 236 · Chapter 7 Specifications and Measurement Uncertainti…
Agilent Technologies 8753D — Chapter 7 Specifications and Measurement Uncertainties diagram (page 456)
Fig. 237 · Chapter 7 Specifications and Measurement Uncertainti…

Speci cations and Measurement Uncertainties 7-25

Page 457 of the Agilent Technologies 8753D user manual — Chapter 7 Specifications and Measurement Uncertainties
Page 457 · click to enlarge
Page 457

The speci cations listed in T able 1 range from those guaranteed by Hewlett-P ackard to those

typical of most HP 8753D instruments , but not guaranteed. Codes in the far right column

of T able 1 reference a speci cation de nition, listed below . These de nitions are intended

to clarify the extent to which Hewlett-P ackard supports the speci ed performance of the

HP 8753D .

S-1This performance parameter is veri able using performance tests documented in the

service manual.

S-2Due to limitations on available industry standards , the guaranteed performance of the

instrument cannot be veri ed outside the factory . Field procedures can verify performance

with a con dence prescribed by available standards .

S-3These speci cations are generally digital functions or are mathematically derived from

tested speci cations , and can therefore be veri ed by functional pass/fail testing.

T: Typical but non-warranted performance characteristics intended to provide information

useful in applying the instrument. Typical characteristics are representative of most

instruments , though not necessarily tested in each unit. Not eld tested.

7-26 Speci cations and Measurement Uncertainties

Instrument Speci cations

Page 458 of the Agilent Technologies 8753D user manual — SPECTRAL PURITY CHARA CTERISTICS
Page 458 · click to enlarge
Page 458

T able 7-12. HP 8753D Instrument Speci cations (1 of 6)

TEST PORT OUTPUT

Description Speci cation Code

FREQUENCY CHARA CTERISTICS

Standard 30 kHz to 3 GHz S-1

Option 006 30 kHz to 6 GHz S-1

A ccuracy (at 25

C 5

C) 10 ppm S-1

C 7.5 ppm T

per year 3 ppm T

Resolution 1 Hz S-3

OUTPUT PO WER CHARA CTERISTICS

Range:

Standard 85 to +10 dBm S-1

Option 075 85 to +8 dBm S-1

Resolution 0.05 dB S-3

Level A ccuracy (at 0 dBm output level) (at 25

C 5

C)

1.0 dB S-1*

C 5

C)

15 to +5 dBm 0.2 dB (relative to 0 dBm output level) S-1

+5 to +10 dBm (Standard) 0.5 dB (relative to 0 dBm output level) S-1

+5 to +8 dBm (Option 075) 0.5 dB (relative to 0 dBm output level) S-1

Standard 50 ohms> 16 dB return loss to 3 GHz T

>14 dB return loss to 6 GHz T

Option 075 75 ohms> 10 dB return loss to 300 kHz T

>16 dB return loss to 3 GHz T

SPECTRAL PURITY CHARA CTERISTICS

2nd Harmonic (16 MHz to 3 GHz)

at +10 dBm output level

at 0 dBm output level

at 10 dBm output level

3rd Harmonic (16 MHz to 2 GHz)

at +10 dBm output level

at 0 dBm output level

at 10 dBm output level

at +10 dBm output level

at 10 dBm output level

* Explicitly tested as part of an on-site veri cation performed by Hewlett-P ackard.

Typical 30 kHz to 300 kHz and typical from 2 to 3 GHz for Option 075.

Speci cations and Measurement Uncertainties 7-27

Show 19 more labels / short text on this page
Range · · · Stability · · to 55 · · · · y · y · Linearity (at 25 · · · Impedance · Impedance · Non-Harmonic Spurious Signals · Mixer Related · y
Page 459 of the Agilent Technologies 8753D user manual — C)
Page 459 · click to enlarge
Page 459

T able 7-12. HP 8753D Instrument Speci cations (2 of 6)

TEST PORT INPUTS

Description Speci cation Code

CHARA CTERISTICS

Standard 30 kHz to 3 GHz S-1

Option 006 30 kHz to 6 GHz S-1

Impedance 50 ohms nominal (Standard)

Standard:

30 kHz to 50 kHz 10 dB return loss T

50 kHz to 300 kHz 18 dB return loss T

300 kHz to 1.3 GHz 18 dB return loss S-1

1.3 GHz to 3 GHz 16 dB return loss S-1

3 GHz to 6 GHz 14 dB return loss S-1

Impedance 75 ohms nominal (Option 075)

Option 075:

30 kHz to 300 kHz 10 dB return loss T

300 kHz to 3 GHz 16 dB return loss T

Maximum Input Level +10 dBm S-1

Damage Level +26 dBm or > 35 Vdc T

A verage Noise Level

300 kHz to 3 GHz

3 kHz IF bandwidth 82 dBm S-1*

10 Hz IF bandwidth 102 dBm S-1*

110 dBm T

3 GHz to 6 GHz

3 kHz IF bandwidth 77 dBm S-1*

10 Hz IF bandwidth 97 dBm S-1*

105 dBm T

Frequency Response (25 5

C)

300 kHz to 3 GHz 1 dB S-1*

3 GHz to 6 GHz 2 dB S-1*

Internally Generated Harmonics (option 002)

at +8 dBm input level

at +0 dBm input level

at 15 dBm input level

at +8 dBm input level

at +0 dBm input level

at 15 dBm input level

Harmonic Measurement A ccuracy (25 5

C)

16 MHz to 3 GHz 1 dB S-1

3 GHz to 6 GHz

3 dB S-1

Harmonic Measurement Dynamic Range

(with output at 10 dBm and input at < 15 dBm) 40 dBc T

* Explicitly tested as part of an on-site veri cation performed by Hewlett-P ackard.

Operation from 3 GHz to 6 GHz requires Option 006.

7-28 Speci cations and Measurement Uncertainties

Show 7 more labels / short text on this page
Frequency Range · · 2nd Harmonic · 3rd Harmonic · · y · y
Page 460 of the Agilent Technologies 8753D user manual — R CHANNEL INPUT
Page 460 · click to enlarge
Page 460

T able 7-12. HP 8753D Instrument Speci cations (3 of 6)

R CHANNEL INPUT

Description Speci cation Code

Frequency O set Operation *

300 kHz to 6 GHz S-1

R Channel Input Requirements 0 to 35 dBm, to 3 GHz S-1

(required for phase-locked operation) 0 to 30 dBm, 3 GHz to 6 GHz S-1

0 to 30 dBm, 3 GHz to 6 GHz S-1

LO Spectral Purity and A ccuracy

Maximum Spurious Input

Residual FM

Frequency A ccuracy 1 to +1 MHz of nominal frequency T

A ccuracy (see Magnitude Characteristics and Phase Characteristics )

(CW Time sweep only)

300 kHz to 6 GHz S-1

P ower Level 0 to 25 dBm T

Maximum Spurious Input

Residual FM

A uto 500 ms T

Manual 50 ms T

Frequency Readout A ccuracy (auto) 0.1% T

Manual 0.5 to 5 MHz T

50 MHz 5 MHz of nominal CW frequency T

>50 MHz 10% of nominal CW frequency T

A ccuracy (see Magnitude Characteristics and Phase Characteristics )

* The HP 8753D RF source characteristics in this mode are dependent on the stability of the external LO source .

The RF source tracks the LO to maintain a stable IF signal at the R channel receiver input. Degradation in

accuracy is negligible with an HP 8642A/B or HP 8656B RF signal generator as the LO source .

Refer to \HP 8753D Descriptions and Options" for a functional description.

Operation from 3 GHz to 6 GHz requires option 006.

Measurement accuracy is dependent on the stability of the input signal.

Speci cations and Measurement Uncertainties 7-29

Show 14 more labels / short text on this page
y · Frequency Range · z · y x · External Source Mode · Frequency Range · z · R Input Requirements · Spectral Purity · Setting Time · Input Frequency Margin · A uto · y · z
Page 461 of the Agilent Technologies 8753D user manual — S-3
Page 461 · click to enlarge
Page 461

T able 7-12. HP 8753D Instrument Speci cations (4 of 6)

INPUT GENERAL

Description Speci cation Code

MA GNITUDE CHARA CTERISTICS

Display Resolution 0.01 dB/division S-3

Resolution 0.001 dB S-3

(10 Hz BW , inputs T est P ort 1 and T est P ort 2; R to 35 dBm) (see graph) S-1

Agilent Technologies 8753D — MA GNITUDE CHARA CTERISTICS diagram (page 461)
Fig. 238

Dynamic A ccuracy (Magnitude)

30 kHz to 3GHz

3 GHz to 6 GHz

Range 500 dB S-3

Resolution 0.001 dB S-3

30 kHz to 3 GHz 0.02 dB/

C T

3 GHz to 6 GHz 0.04 dB/

C T

PHASE CHARA CTERISTICS

S-3

Display Resolution 0.01

S-3

* Marker resolution for magnitude , phase , and delay is dependent upon the value measured; resolution is limited

to 5 digits .

CW sweep , +5 dBm into T est P ort, ratio measurement, 3 kHz BW

7-30 Speci cations and Measurement Uncertainties

Show 17 more labels / short text on this page
Marker · * · Dynamic A ccuracy · Trace Noise · y · Reference Level · Stability · · · Range 180 · · · /division S-3 · * · · Marker Resolution · y
Page 462 of the Agilent Technologies 8753D user manual — S-3
Page 462 · click to enlarge
Page 462

T able 7-12. HP 8753D Instrument Speci cations (5 of 6)

INPUT GENERAL (cont.)

Description Speci cation Code

PHASE CHARA CTERISTICS (cont.)

(10 Hz BW , inputs T est P ort 1 and T est P ort 2; R to 35 dBm) (see graph) S-1

Agilent Technologies 8753D — Chapter 7 Specifications and Measurement Uncertainties diagram (page 462)
Fig. 239 · Chapter 7 Specifications and Measurement Uncertainti…

Dynamic A ccuracy (Phase)

Trace Noise (+5 dBm into T est P ort, ratio measurement)

30 kHz to 3 GHz

3 GHz to 6 GHz

S-3

Resolution 0.01

S-3

30 kHz to 3 GHz 0.05

3 GHz to 6 GHz 0.20

POLAR CHARA CTERISTICS (ratio measurement)

Range 10 10

up to 1000 units full scale S-3

Reference range of 500 units S-3

Speci cations and Measurement Uncertainties 7-31

Show 15 more labels / short text on this page
Dynamic A ccuracy · · rms S-1 · · rms S-1 · Reference Level · Range 500 · · · Stability · · /degree C T · · /degree C T · 12
Page 463 of the Agilent Technologies 8753D user manual — GROUP DELA Y CHARA CTERISTICS
Page 463 · click to enlarge
Page 463

T able 7-12. HP 8753D Instrument Speci cations (6 of 6)

INPUT GENERAL (cont.)

Description Speci cation Code

GROUP DELA Y CHARA CTERISTICS

Group delay is computed by measuring the phase change within

a speci ed frequency step (determined by the frequency span

and the number of points per sweep).

Aperture (selectable) (frequency span)/(number of points 1) S-3

Maximum Aperture 20% of frequency span S-3

Range 1/2 (1/minimum aperture) S-3

(The maximum delay is limited to measuring no more than 180

of phase change within the minimum aperture .)

A ccuracy (see graph) S-3

The following graph shows group delay accuracy with an

HP 85047A T est Set with 7 mm full 2-port calibration and a 10

Hz IF bandwidth. Insertion loss is assumed to be

electrical length to be ten meters .

Agilent Technologies 8753D — GROUP DELA Y CHARA CTERISTICS diagram (page 463)
Fig. 240

Group Delay A ccuracy vs . Aperture

In general, the following formula can be used to determine the accuracy , in seconds , of speci c group delay

measurement:

[0.003 Phase A ccuracy (deg)]/Aperture (Hz)

Depending on the aperture and device length, the phase accuracy used is either incremental phase accuracy or

worst case phase accuracy .

frequency span and the number of points per sweep).

Aperture (selectable) (frequency span)/(number of points 1) S-3

Maximum Aperture 20% of frequency span S-3

Range 1/2 (1/minimum aperture) S-3

(The maximum delay is limited to measuring no more than 180

of phase change within the minimum aperture .)

A ccuracy (see graph) S-3

7-32 Speci cations and Measurement Uncertainties

·

Page 464 of the Agilent Technologies 8753D user manual — ASCII
Page 464 · click to enlarge
Page 464

HP 8753D Network Analyzer General Characteristics

The following table shows typical measurement times for the HP 8753D network analyzer in

milliseconds .

T able 7-13. Typical Measurement Times

Typical Time for Completion (ms)

Number of P oints

51 201 401 1601

Uncorrected 125 200 300 900

125 200 300 900

245 510 855 2940

80 350 740 1790

Binary (Internal) 20 35 55 205

IEEE754 oating point format

32 bit 25 85 150 590

64 bit 40 115 220 840

ASCII 140 510 1000 3960

S11 1-port calibration, with a 3 kHz IF bandwidth. Includes system retrace time , but does not include

bandswitch time . Time domain gating is assumed o .

S21 measurement with full 2-port calibration, using a 3 kHz IF bandwidth. Includes system retrace time and

RF switching time , but does not include bandswitch time . Time domain gating is assumed o .

Option 010 only , gating o .

Measured with HP 9000 series 300 computer .

HP-IB interface operates according to IEEE 488-1978 and IEC 625 standards and IEEE 728-1982

recommended practices .

Binary (internal 48-bit oating point complex format)

ASCII

32/64 bit IEEE 754 Floating P oint F ormat

SH1, AH1, T6, TE0, L4, LE0, SR1, RL1, PP0, DC1, DT1, C1, C2, C3, C10, E2

Speci cations and Measurement Uncertainties 7-33

Show 16 more labels / short text on this page
Measurement Throughput Summary · Measurement · 1-port cal · * · 2-port cal · y · z · Time Domain Conversion · HP-IB Data Transfer · * · y · z · Remote Programming · Interface · Transfer F ormats · Interface Function Codes
Page 465 of the Agilent Technologies 8753D user manual — Chapter 7 Specifications and Measurement Uncertainties
Page 465 · click to enlarge
Page 465

Front P anel Connectors

Connector Type : 7 mm precision

Impedance : 50 ohms (nominal)

Connector Conductor Depth : 0.000 to 0.003 in.

+15 V 2% 400 mA (combined load for both probe connections)

12.6 V 5.5% 300 mA (combined load for both probe connections)

Rear P anel Connectors

External Reference Frequency Input (EXT REF INPUT)

Frequency : 1, 2, 5, and 10 MHz (200 Hz at 10 MHz)

Level : 10 dBm to +20 dBm, typical

Impedance : 50 ohms

High-Stability Frequency Reference Output (10 MHz)(Option 001)

Frequency : 10.0000 MHz

C) : 0.05 ppm

Daily Aging Rate (after 30 days) : 3 10

Y early Aging Rate : 0.5 ppm/year

Output : 0 dBm minimum

Nominal Output Impedance : 50

External A uxiliary Input (A UX INPUT)

Input V oltage Limits : 10 V to +10 V

External AM Input (EXT AM)

1 volt into a 5 k resistor , 1 kHz maximum, resulting in approximately 8 dB/volt amplitude

modulation.

7-34 Speci cations and Measurement Uncertainties

Show 7 more labels / short text on this page
Probe P ower · Frequency Stability (0 · · C to 55 · · 9 · /day
Page 466 of the Agilent Technologies 8753D user manual — HP-IB
Page 466 · click to enlarge
Page 466

External Trigger (EXT TRIGGER)

Triggers on a negative TTL transition or contact closure to ground.

Agilent Technologies 8753D — Chapter 7 Specifications and Measurement Uncertainties diagram (page 466)
Fig. 241 · Chapter 7 Specifications and Measurement Uncertainti…

Figure 7-1. External Trigger Circuit

T est Sequence Output (TEST SEQ)

This connector outputs a TTL signal which can be programmed by the user in a test sequence

to be high or low . By default, this output provides an end-of-sweep TTL signal. (F or use with

part handlers .)

Limit T est Output (LIMIT TEST)

This connector outputs a TTL signal of the limit test results . P ass: TTL high; F ail: TTL low .

T est P ort Bias Input (BIAS CONNECT)

Maximum voltage : +30 Vdc

Maximum current (no degradation in RF speci cations) : 200 mA

Maximum current : 1 A

Video Output (EXT MON)

The R, G, and B connectors drive external monitors with these characteristics:

R, G, B with synch on green.

75 ohm impedance .

1 Vp-p (0.7 V=white; 0 V=black; 0.3 V=synch).

HP-IB

This connector allows communication with compatible devices including external controllers ,

printers , plotters , disk drives , and power meters .

Speci cations and Measurement Uncertainties 7-35

Page 467 of the Agilent Technologies 8753D user manual — RS-232
Page 467 · click to enlarge
Page 467

P arallel P ort

This connector is used with parallel (or Centronics interface) peripherals such as printers

and plotters . It can also be used as a general purpose I/O port, with control provided by test

sequencing functions .

RS-232

This connector is used with serial peripherals such as printers and plotters .

This connector is used for the optional A T compatible keyboard for titles and remote

front-panel operation.

48 to 66 Hz

115 V nominal (90 V to 132 V) or 230 V nominal (198 V to 264 V). 280 V A max.

RFI and EMI susceptibilityde ned by V ADE 0730, CISPR Publication 11, and FCC Class B

Standards .

ESD (electrostatic discharge)must be eliminated by use of static-safe work procedures and

an anti-static bench mat (such as HP 92175T).

Dust: the environment should be as dust-free as possible .

Operating T emperature : 0

Error-Corrected T emperature Range : 1

C of calibration temperature

Humidity : 5% to 95% at 40

Altitude : 0 to 4500 meters (15,000 feet)

T emperature : 40

Humidity : 0 to 90% relative at +65

Altitude : 0 to 15,240 meters (50,000 feet)

7-36 Speci cations and Measurement Uncertainties

Show 17 more labels / short text on this page
DIN K eyboard · Line P ower · Environmental Characteristics · General Conditions · Operating Conditions · · to 55 · · · · C (non-condensing) · Non-Operating Storage Conditions · · C to +70 · · · C (non-condensing)
Page 468 of the Agilent Technologies 8753D user manual — Chapter 7 Specifications and Measurement Uncertainties
Page 468 · click to enlarge
Page 468

Net : 34 kg (75 lb)

Shipping : 37 kg (82 lb)

222 mm H 425 mm W 508 mm D

(8.75 16.75 20.0 in)

(These dimensions exclude front and rear panel protrusions .)

Agilent Technologies 8753D — Chapter 7 Specifications and Measurement Uncertainties diagram (page 468)
Fig. 242 · Chapter 7 Specifications and Measurement Uncertainti…

Data Retention Time with 3 V , 1.2 Ah Battery

T emperature at 70

C : 250 days (0.68 year) characteristically

T emperature at 40

C : 1244 days (3.4 years) characteristically

T emperature at 25

C : 10 years characteristically

Speci cations and Measurement Uncertainties 7-37

Show 7 more labels / short text on this page
W eight · Cabinet Dimensions · Physical Dimensions · Internal Memory · · ·
Page 469 of the Agilent Technologies 8753D user manual — Chapter 7 Specifications and Measurement Uncertainties
Page 469 · click to enlarge
Page 469
Page 469 of the Agilent Technologies 8753D user manual — Chapter 7 Specifications and Measurement Uncertainties
Page 470 of the Agilent Technologies 8753D user manual — A VG
Page 470 · click to enlarge
Page 470

This chapter contains menus maps arranged in the following order:

A VG

CAL

COPY

DISPLA Y

F ORMA T

LOCAL

MARKER

MARKER F CTN

MEAS

MENU

SA VE/RECALL

PRESET

SCALE REF

SEQ

SYSTEM

Show 32 more labels / short text on this page
Menu Maps · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · Menu Maps 8-1
Page 471 of the Agilent Technologies 8753D user manual — Chapter 8 Menu Maps
Page 471 · click to enlarge
Page 471
Agilent Technologies 8753D — Chapter 8 Menu Maps diagram (page 471)
Fig. 243 · Chapter 8 Menu Maps

d a · c b · 8-2 Menu Maps

Page 472 of the Agilent Technologies 8753D user manual — Chapter 8 Menu Maps
Page 472 · click to enlarge
Page 472
Agilent Technologies 8753D — Chapter 8 Menu Maps diagram (page 472)
Fig. 244 · Chapter 8 Menu Maps
Page 473 of the Agilent Technologies 8753D user manual — Chapter 8 Menu Maps
Page 473 · click to enlarge
Page 473
Agilent Technologies 8753D — Chapter 8 Menu Maps diagram (page 473)
Fig. 245 · Chapter 8 Menu Maps
Page 474 of the Agilent Technologies 8753D user manual — Chapter 8 Menu Maps
Page 474 · click to enlarge
Page 474
Agilent Technologies 8753D — Chapter 8 Menu Maps diagram (page 474)
Fig. 246 · Chapter 8 Menu Maps

d a · c b · Menu Maps 8-5

Page 475 of the Agilent Technologies 8753D user manual — Chapter 8 Menu Maps
Page 475 · click to enlarge
Page 475
Agilent Technologies 8753D — Chapter 8 Menu Maps diagram (page 475)
Fig. 247 · Chapter 8 Menu Maps
Agilent Technologies 8753D — Chapter 8 Menu Maps diagram (page 475)
Fig. 248 · Chapter 8 Menu Maps
Show 5 more labels / short text on this page
d a · c b · d a · c b · 8-6 Menu Maps
Page 476 of the Agilent Technologies 8753D user manual — Chapter 8 Menu Maps
Page 476 · click to enlarge
Page 476
Agilent Technologies 8753D — Chapter 8 Menu Maps diagram (page 476)
Fig. 249 · Chapter 8 Menu Maps

d a · c b · Menu Maps 8-7

Page 477 of the Agilent Technologies 8753D user manual — Chapter 8 Menu Maps
Page 477 · click to enlarge
Page 477
Agilent Technologies 8753D — Chapter 8 Menu Maps diagram (page 477)
Fig. 250 · Chapter 8 Menu Maps

d a · c b · 8-8 Menu Maps

Page 478 of the Agilent Technologies 8753D user manual — Chapter 8 Menu Maps
Page 478 · click to enlarge
Page 478
Agilent Technologies 8753D — Chapter 8 Menu Maps diagram (page 478)
Fig. 251 · Chapter 8 Menu Maps

d a · c b · Menu Maps 8-9

Page 479 of the Agilent Technologies 8753D user manual — Chapter 8 Menu Maps
Page 479 · click to enlarge
Page 479
Agilent Technologies 8753D — Chapter 8 Menu Maps diagram (page 479)
Fig. 252 · Chapter 8 Menu Maps

d a · c b · 8-10 Menu Maps

Page 480 of the Agilent Technologies 8753D user manual — Chapter 8 Menu Maps
Page 480 · click to enlarge
Page 480
Agilent Technologies 8753D — Chapter 8 Menu Maps diagram (page 480)
Fig. 253 · Chapter 8 Menu Maps

d a · c b · Menu Maps 8-11

Page 481 of the Agilent Technologies 8753D user manual — Chapter 8 Menu Maps
Page 481 · click to enlarge
Page 481
Agilent Technologies 8753D — Chapter 8 Menu Maps diagram (page 481)
Fig. 254 · Chapter 8 Menu Maps
Agilent Technologies 8753D — Chapter 8 Menu Maps diagram (page 481)
Fig. 255 · Chapter 8 Menu Maps
Show 5 more labels / short text on this page
d a · c b · d a · c b · 8-12 Menu Maps
Page 482 of the Agilent Technologies 8753D user manual — Chapter 8 Menu Maps
Page 482 · click to enlarge
Page 482
Agilent Technologies 8753D — Chapter 8 Menu Maps diagram (page 482)
Fig. 256 · Chapter 8 Menu Maps
Page 483 of the Agilent Technologies 8753D user manual — Chapter 8 Menu Maps
Page 483 · click to enlarge
Page 483
Agilent Technologies 8753D — Chapter 8 Menu Maps diagram (page 483)
Fig. 257 · Chapter 8 Menu Maps
Page 484 of the Agilent Technologies 8753D user manual — Chapter 8 Menu Maps
Page 484 · click to enlarge
Page 484

8-14 Menu Maps

Page 485 of the Agilent Technologies 8753D user manual — Chapter 8 Menu Maps
Page 485 · click to enlarge
Page 485
Agilent Technologies 8753D — Chapter 8 Menu Maps diagram (page 485)
Fig. 258 · Chapter 8 Menu Maps

d a · c b · Menu Maps 8-15

Page 486 of the Agilent Technologies 8753D user manual — Chapter 8 Menu Maps
Page 486 · click to enlarge
Page 486
Page 486 of the Agilent Technologies 8753D user manual — Chapter 8 Menu Maps
Page 487 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 487 · click to enlarge
Page 487

K ey De nitions

This chapter contains information on the following topics:

Softkey and front-panel functions in alphabetical order (includes a brief description of each

Cross reference of programming commands to key functions

Cross reference of softkeys to front-panel access keys

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SERVICE MENU keys are not included in this chapter . Service information

can be found in the HP 8753D Network Analyzer Service Guide .

Where to Look for More Information

A dditional information about many of the topics discussed in this chapter is located in the

following areas:

Chapter 2, \Making Measurements ," contains step-by-step procedures for making

measurements or using particular functions .

Chapter 4, \Printing, Plotting, and Saving Measurement Results ," contains instructions

for saving to disk or the analyzer internal memory , and printing and plotting displayed

measurements .

Chapter 5, \Optimizing Measurement Results ," describes techniques and functions for

achieving the best measurement results .

Chapter 6, \Application and Operation Concepts ," contains explanatory-style information

about many applications and analyzer operation.

HP 8753D Network Analyzer Programmer's Guide provides a complete description of all

HP-IB mnemonics .

K ey De nitions 9-1

function) · The

Page 488 of the Agilent Technologies 8753D user manual —  REF = 2
Page 488 · click to enlarge
Page 488

Guide T erms and Conventions

The eight keys along the right side of the analyzer display are called softkeys . Their labels

are shown on the display . The softkeys appear in shaded boxes in this chapter . F or example ,

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRANSMISSION . The labeled keys that are on the front panel of the analyzer are called

front-panel keys . The front-panel keys appear in unshaded boxes in this chapter . F or example ,

SYSTEM

.

This section contains an alphabetical listing of softkey and front-panel functions , and a brief

description of each function.

is used to add a decimal point to the number you are entering.

is used to add a minus sign to the number you are entering.

is used to step up the current value of the active function.

The analyzer de nes the step for di erent functions . No units

terminator is required. F or editing a test sequence , this key can

be used to scroll through the displayed sequence .

is used to step down the current value of the active function.

The analyzer de nes the step for di erent functions . No units

terminator is required. F or editing a test sequence , this key can

be used to scroll through the displayed sequence .

is used to delete the last entry , or the last digit entered from

the numeric keypad. This key can also be used in two ways for

modifying a test sequence:

deleting a single-key command that you may have pressed by

NNNNNNNNNNN

mistake , (for example

A/R )

deleting the last digit in a series of entered digits , as long as

you haven't yet pressed a terminator , (for example if you

ST ART

2 but did not press

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

goes to the delta marker menu, which is used to read the

MODE MENU

di erence in values between the active marker and a reference

marker .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MODE OFF turns o the delta marker mode , so that the values displayed

for the active marker are absolute values .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REF = 1 establishes marker 1 as a reference . The active marker stimulus

and response values are then shown relative to this delta

reference . Once marker 1 has been selected as the delta

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

reference , the softkey label

REF = 1 is underlined in this

menu, and the marker menu is returned to the screen. In the

marker menu, the rst key is now labeled

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER REF = 1 .

The notation \REF=1" appears at the top right corner of the

graticule .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

makes marker 2 the delta reference . A ctive marker stimulus

REF = 2

and response values are then shown relative to this reference .

9-2 K ey De nitions

Show 15 more labels / short text on this page
· Analyzer Functions · · · · · · · · · pressed · · 1 · G/n · , etc)
Page 489 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 489 · click to enlarge
Page 489

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REF = 3

makes marker 3 the delta reference .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

makes marker 4 the delta reference .

REF = 4

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

makes marker 5 the delta reference .

REF = 5

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REF = FIXED MKR

sets a user-speci ed xed reference marker . The stimulus and

response values of the reference can be set arbitrarily , and can

be anywhere in the display area. Unlike markers 1 to 5, the

xed marker need not be on the trace . The xed marker is

indicated by a small triangle , and the active marker stimulus

and response values are shown relative to this point. The

graticule .

Pressing this softkey turns on the xed marker . Its stimulus and

response values can then be changed using the xed marker

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

menu, which is accessed with the

FIXED MKR POSITION

softkey described below . Alternatively , the xed marker can be

set to the current active marker position, using the

NNNNNNNNNNNNNNNNNNNNNNNNNN

MKR ZERO

softkey in the marker menu.

NNNNNNNNNNN

expresses the data in inverse S-parameter values , for use in

1/S

ampli er and oscillator design.

NNNNN

measures the absolute power amplitude at input A.

A

NNNNNNNNNNN

A/B

calculates and displays the complex ratio of input A to input B .

NNNNNNNNNNN

calculates and displays the complex ratio of the signal at input

A/R

A to the reference signal at input R.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

puts the name of the active entry in the display title .

ACTIVE ENTRY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ACTIVE MRK MAGNITUDE

puts the active marker magnitude in the display title .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

selects coaxial as the type of port used in adapter removal

ADAPTER: COAX

calibration.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ADAPTERWAVEGUIDE selects waveguide as the type of port used in adapter removal

calibration.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ADAPTER DELAY is used to enter the value of electrical delay of the adapter used

in adapter removal calibration.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ADAPTER REMOVAL

provides access to the adapter removal menu.

NNNNNNNNNNN

displays the edit segment menu and adds a new segment to

ADD

the end of the list. The new segment is initially a duplicate of

the segment indicated by the pointer > and selected with the

NNNNNNNNNNNNNNNNNNNNNNN

SEGMENT softkey .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

sets the HP-IB address of the analyzer , using the entry controls .

ADDRESS: 8753

There is no physical address switch to set in the analyzer .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ADDRESS: CONTROLLER

sets the HP-IB address the analyzer will use to communicate

with the external controller .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ADDRESSDISK sets the HP-IB address the analyzer will use to communicate

with an external HP-IB disk drive .

K ey De nitions 9-3

Page 490 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 490 · click to enlarge
Page 490

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ADDRESSP MTR/HPIB sets the HP-IB address the analyzer will use to communicate

with the power meter used in service routines .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

presents a menu for adjusting display intensity , colors , and

ADJUST DISPLAY

accessing save and recall functions for modi ed LCD color sets .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

leads to the beginning of the adjustment tests . These tests

ADJUSTMENT TESTS

generate correction constants that are used by the analyzer .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

retrieves the full frequency list sweep .

ALL SEGS SWEEP

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ALTERNATE A and B

measures only one input per frequency sweep , in order to

reduce spurious signals . Thus , this mode optimizes the dynamic

range for all four S-parameter measurements .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AMPLITUDE OFFSET

adds or subtracts an o set in amplitude value . This allows

limits already de ned to be used for testing at a di erent

response level. F or example , if attenuation is added to or

removed from a test setup , the limits can be o set an equal

amount. Use the entry block controls to specify the o set.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ANALOG IN Aux Input

displays a dc or low frequency ac auxiliary voltage on the

vertical axis , using the real format. An external signal source

such as a detector or function generator can be connected to

the rear panel A UXILIARY INPUT connector .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ASSERT SRQ

sets the sequence bit in the Event Status Register , which can

be used to generate an SRQ (service request) to the system

controller .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AUTO FEED ON off turns the plotter auto feed function on or o when in the

de ne plot menu. It turns the printer auto feed on or o when

in the de ne print menu.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AUTO SCALE

brings the trace data in view on the display with one keystroke .

Stimulus values are not a ected, only scale and reference

values . The analyzer determines the smallest possible scale

factor that will put all displayed data onto 80% of the vertical

graticule . The reference value is chosen to put the trace in

center screen, then rounded to an integer multiple of the scale

factor .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AUX OUT on OFF allows you to monitor the analog bus nodes (except nodes

1,2,3,4,9,10,12) with external equipment. T o do this , connect

the equipment to the A UX INPUT BNC connector on the rear

panel.

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

makes averaging factor the active function. Any value up to

AVERAGING FACTOR

999 can be used. The algorithm used for averaging is:

A(n) = S (n)=F + (1 1=F ) A(n 1)

A(n) = current average

S(n) = current measurement

F = average factor

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

turns the averaging function on or o for the active channel.

\A vg" is displayed in the status notations area at the left of the

display , together with the sweep count for the averaging factor ,

9-4 K ey De nitions

where · AVERAGING on OFF

Page 491 of the Agilent Technologies 8753D user manual — NNNNNNNN
Page 491 · click to enlarge
Page 491

when averaging is on. The sweep count for averaging is reset to

1 whenever an instrument state change a ecting the measured

data is made .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

At the start of the averaging or following

AVERAGING RESTART ,

averaging starts at 1 and averages each new sweep into the

trace until it reaches the speci ed averaging factor . The sweep

count is displayed in the status notations area below \A vg" and

updated every sweep as it increments . When the speci ed

averaging factor is reached, the trace data continues to be

updated, weighted by that averaging factor .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AVERAGING RESTART

averaging starts at 1 and averages each new sweep into the

trace until it reaches the speci ed averaging factor . The sweep

count is displayed in the status notations area below \A vg" and

updated every sweep as it increments .

A VG

is used to access three di erent noise reduction techniques:

sweep-to-sweep averaging, display smoothing, and variable IF

bandwidth. Any or all of these can be used simultaneously .

A veraging and smoothing can be set independently for each

channel, and the IF bandwidth can be set independently if the

stimulus is uncoupled.

NNNNN

B

measures the absolute power amplitude at input B .

NNNNNNNNNNN

calculates and displays the complex ratio of input B to input R.

B/R

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

BACK SPACE

deletes the last character entered.

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

sets the background intensity of the LCD as a percent of white .

BACKGROUND INTENSITY

The factory-set default value is stored in non-volatile memory .

NNNNNNNNNNNNNNNNNNNNNNNNNN

BANDPASS

(Option 010 only) sets the time-domain bandpass mode .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

toggles an annunciator which sounds to indicate completion of

BEEP DONE ON off

certain operations such as calibration or instrument state save .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

turns the limit fail beeper on or o . When limit testing is on

BEEP FAIL on OFF

and the fail beeper is on, a beep is sounded each time a limit

test is performed and a failure detected. The limit fail beeper is

independent of the warning beeper and the operation complete

beeper .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

toggles the warning annunciator . When the annunciator is on it

BEEP WARN on OFF

sounds a warning when a cautionary message is displayed.

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

switches o the analyzer's display . This feature may be helpful

BLANK DISPLAY

in prolonging the life of the LCD in applications where the

analyzer is left unattended (such as in an automated test

system). Pressing any front panel key will restore the default

display operation.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

adjusts the brightness of the color being modi ed. See A djusting

BRIGHTNESS

Color for an explanation of using this softkey for color

modi cation of display attributes .

NNNNNNNN

is used to enter the C0 term in the de nition of an OPEN

standard in a calibration kit, which is the constant term of the

cubic polynomial and is scaled by 10

K ey De nitions 9-5

· 15

Page 492 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 492 · click to enlarge
Page 492

NNNNNNNN

is used to enter the C1 term, expressed in F/Hz (F arads/Hz) and

N

NNNNNNN

is used to enter the C2 term, expressed in F/Hz

NNNNNNNN

C3 is used to enter the C3 term, expressed in F/Hz

key leads to a series of menus to perform measurement

CAL

calibrations for vector error correction (accuracy enhancement),

N

NNNNNNNNNN

and for specifying the calibration standards used. The

CAL

key also leads to softkeys which activate interpolated error

correction and power meter calibration.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL FACTOR

accepts a power sensor calibration factor % for the segment.

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

brings up the segment modify menu and segment edit

CAL FACTOR SENSOR A

(calibration factor menu) which allows you to enter a power

sensor's calibration factors . The calibration factor data entered

in this menu will be stored for power sensor A.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL FACTOR SENSOR B

brings up the segment modify menu and segment edit

(calibration factor menu) which allows you to enter a power

sensor's calibration factors . The calibration factor data entered

in this menu will be stored for power sensor B .

NNNNNNNNNNNNNNNNNNNNNNN

CAL KIT

leads to the select cal kit menu, which is used to select one of

the default calibration kits available for di erent connector

types . This , in turn, leads to additional menus used to de ne

calibration standards other than those in the default kits (refer

to Modifying Calibration Kits .) When a calibration kit has been

speci ed, its connector type is displayed in brackets in the

softkey label.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL KIT [ ]

selects the cal kit menu.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

selects the HP 85056A/D cal kit.

CAL KIT2.4mm

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

selects the HP 85056K cal kit.

CAL KIT: 2.92*

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL KIT2.92mm

selects the 2.92 mm cal kit model.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL KIT3.5mmC

selects the HP 85033C cal kit.

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

selects the HP 85033D cal kit.

CAL KIT3.5mmD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

selects the HP 85052C TRL cal kit.

CAL KITTRL 3.5mm

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL KIT7mm

selects the HP 85031B cal kit.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

selects the HP 85032B cal kit.

CAL KIT: N 50

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

selects the HP 85036B/E cal kit.

CAL KIT: N 75

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL KIT: USER KIT

selects a kit other than those o ered by Hewlett-P ackard.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

this default selection establishes the TRL/LRM LINE/MA TCH

CAL ZO: LINE ZO

standard as the characteristic impedance .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL ZO: SYSTEM ZO

allows you to modify the characteristic impedance of the

system for TRL/LRM calibration.

9-6 K ey De nitions

Show 7 more labels / short text on this page
scaled by 10 · 27 · and scaled by · 36 · and scaled by · 45 ·
Page 493 of the Agilent Technologies 8753D user manual — N
Page 493 · click to enlarge
Page 493

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CALIBRATE MENU

leads to the calibration menu, which provides several accuracy

enhancement procedures ranging from a simple frequency

response calibration to a full two-port calibration. At the

completion of a calibration procedure , this menu is returned

to the screen, correction is automatically turned on, and the

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CALIBRATE: NONE

is underlined if no calibration has been performed or if the

calibration data has been cleared. Unless a calibration is saved

in memory , the calibration data is lost on instrument preset,

power on, instrument state recall, or if stimulus values are

changed.

CENTER

is used, along with the

SP AN

key , to de ne the frequency range

of the stimulus . When the

CENTER

key is pressed, its function

becomes the active function. The value is displayed in the

active entry area, and can be changed with the knob , step keys ,

or numeric keypad.

NNNNNNNNNNNNNNNNNNNN

CENTER

sets the center frequency of a subsweep in a list frequency

sweep .

CHAN 1 allows you to select channel 1 as the active channel. The

active channel is indicated by an amber LED adjacent to the

corresponding channel key . The front panel keys allow you

to control the active channel, and all of the channel-speci c

functions you select apply to the active channel.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CH1 DATA [ ]

brings up the printer color selection menu. The channel 1 data

trace default color is magenta for color prints .

N

CH1 DATA LIMIT LN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

selects channel 1 data trace and limit line for display color

modi cation.

NNNNNNNNNNNNNNNNNNNNNNN

CH1 MEM selects channel 1 memory trace for display color modi cation.

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

brings up the printer color selection menu. The channel 1

CH1 MEM [ ]

memory trace default color is green for color prints .

CHAN 2 allows you to select channel 2 as the active channel. The

active channel is indicated by an amber LED adjacent to the

corresponding channel key . The front panel keys allow you

to control the active channel, and all of the channel-speci c

functions you select apply to the active channel.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CH2 DATA [ ]

brings up the printer color selection menu. The channel 2 data

trace default color is blue for color prints .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CH2 DATA LIMIT LN

selects channel 2 data trace and limit line for display color

modi cation.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CH2 MEM [ ] brings up the printer color selection menu. The channel 2

memory trace default color is red for color prints .

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CH2 MEM REF LINE

selects channel 2 memory and the reference line for display

color modi cation.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CH PWR [COUPLED] is used to apply the same power levels to each channel.

N

CH PWR [UNCOUPLED]

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

is used to apply di erent power levels to each channel.

K ey De nitions 9-7

· ·

Page 494 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 494 · click to enlarge
Page 494

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CHOP A and B

measures A and B inputs simultaneously for faster

measurements .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CLEAR BIT when the parallel port is con gured for GPIO , 8 output bits can

be controlled with this key . When this key is pressed, \TTL

OUT BIT NUMBER" becomes the active function. This active

function must be entered through the keypad number keys ,

followed by the

key . The bit is cleared when the

is pressed. Entering numbers larger than 7 will result in bit 7

being cleared, and entering numbers lower than 0 will result in

bit 0 being cleared.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CLEAR LIST

deletes all segments in the list.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

clears a sequence from memory . The titles of cleared sequences

CLEAR SEQUENCE

will remain in load, store , and purge menus . This is done as a

convenience for those who often reuse the same titles .

NNNNNNNNNNNNNN

COAX de nes the standard (and the o set) as coaxial. This causes the

analyzer to assume linear phase response in any o sets .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

applies a linear phase compensation to the trace for use

COAXIAL DELAY

with electrical delay . That is , the e ect is the same as if a

corresponding length of perfect vacuum dielectric coaxial

transmission line was added to the reference signal path.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

provides access to the con gure menu. This menu contains

CONFIGURE

softkeys used to control raw o sets , spur avoidance , and the

test set transfer switch.

NNNNNNNNNNNNNNNNN

COLOR adjusts the degree of whiteness of the color being modi ed. See

\A djusting Color" for an explanation of using this softkey for

color modi cation of display attributes .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

provides access to the con gure ext disk menu. This menu

CONFIGURE EXT DISK

contains softkeys used to the disk address , unit number , and

volume number .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CONTINUE SEQUENCE resumes a paused sequence .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

located under the

key , is the standard sweep mode of

CONTINUOUS

MENU

the analyzer , in which the sweep is triggered automatically and

continuously and the trace is updated with each sweep .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

brings up the conversion menu which converts the measured

CONVERSION [ ]

data to impedance (Z) or admittance (Y). When a conversion

parameter has been de ned, it is shown in brackets under the

softkey label. If no conversion has been de ned, the softkey

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CONVERSION [OFF] .

COPY

provides access to the menus used for controlling external

plotters and printers and de ning the plot parameters .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

turns error correction on or o . The analyzer uses the most

recent calibration data for the displayed parameter . If the

stimulus state has been changed since calibration, the original

state is recalled, and the message \SOURCE P ARAMETERS

CHANGED" is displayed.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

COUNTER: ANALOG BUS

switches the counter to count the analog bus .

9-8 K ey De nitions

Show 4 more labels / short text on this page
key · label reads · · CORRECTION on OFF
Page 495 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 495 · click to enlarge
Page 495

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

COUNTER: DIV FRAC N

switches the counter to count the A14 fractional-N V CO

frequency after it has been divided down to 100 kHz for

phase-locking the V CO .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

COUNTER: FRAC N

switches the counter to count the A14 fractional-N V CO

frequency at the node shown on the overall block diagram.

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

COUNTER: OFF

switches the internal counter o and removes the counter

display from the LCD .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

COUPLED CH on OFF

toggles the channel coupling of stimulus values . With

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

COUPLED CH ON (the preset condition), both channels have

the same stimulus values (the inactive channel takes on the

stimulus values of the active channel).

NNNNNNNNNNNNNNNNNNNNNNN

is used to set the frequency for power sweep and CW time

CW FREQ

sweep modes . If the instrument is not in either of these two

modes , it is automatically switched into CW time mode .

NNNNNNNNNNNNNNNNNNNNNNN

turns on a sweep mode similar to an oscilloscope . The analyzer

CW TIME

is set to a single frequency , and the data is displayed versus

time . The frequency of the CW time sweep is set with

NNNNNNNNNNNNNNNNNNNNNNN

CW FREQ

in the stimulus menu.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

D2/D1 to D2 on OFF this math function ratios channels 1 and 2, and puts the results

in the channel 2 data array . Both channels must be on and

have the same number of points . Refer to Chapter 2, \Making

Measurements" for information on how to use this function to

make gain compression measurements .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

displays both the current data and memory traces .

N

DATA ARRAY on OFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

speci es whether or not to store the error-corrected data on

disk with the instrument state .

NNNNNNNNNNNNNNNNNNNNNNNNNN

DATA/MEM divides the data by the memory , normalizing the data to

the memory , and displays the result. This is useful for ratio

comparison of two traces , for instance in measurements of gain

or attenuation.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DATA - MEM subtracts the memory from the data. The vector subtraction is

performed on the complex data. This is appropriate for storing

a measured vector error , for example directivity , and later

subtracting it from the device measurement.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DATA ! MEMORY

stores the current active measurement data in the memory of

the active channel. It then becomes the memory trace , for use

in subsequent math manipulations or display . If a parameter

has just been changed and the * status notation is displayed at

the left of the display , the data is not stored in memory until

a clean sweep has been executed. The gating and smoothing

status of the trace are stored with the measurement data.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DATA ONLY on OFF stores only the measurement data of the device under test to a

disk le . The instrument state and calibration are not stored.

This is faster than storing with the instrument state , and uses

less disk space . It is intended for use in archiving data that will

later be used with an external controller , and cannot be read

back by the analyzer .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DECISION MAKING

presents the sequencing decision making menu.

K ey De nitions 9-9

DATA and MEMORY

Page 496 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 496 · click to enlarge
Page 496

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DECR LOOP COUNTER

decrements the value of the loop counter by 1.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

returns all the display color settings back to the factory-set

DEFAULT COLORS

default values that are stored in non-volatile memory .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

resets the plotting parameters to their default values .

DEFAULT PLOT SETUP

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DEFAULT PRNT SETUP

resets the printing parameters to their default values .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

leads to the de ne save menu. Use this menu to specify the

DEFINE DISK-SAVE

data to be stored on disk in addition to the instrument state .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DEFINE PLOT leads to a sequence of three menus . The rst de nes which

elements are to be plotted and the auto feed state . The second

de nes which pen number is to be used with each of the

elements (these are channel dependent.) The third de nes the

line types (these are channel dependent), plot scale , and plot

speed.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

leads to the de ne print menu. This menu de nes the printer

DEFINE PRINT

mode (monochrome or color) and the auto-feed state .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

makes the standard number the active function, and brings up

DEFINE STANDARD

the de ne standard menus . The standard number (1 to 8) is an

arbitrary reference number used to reference standards while

specifying a class .

NNNNNNNNNNNNNNNNN

selects the group delay format, with marker values given in

DELAY

seconds .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DELAY/THRU de nes the standard type as a transmission line of speci ed

length, for calibrating transmission measurements .

NNNNNNNNNNNNNNNNNNNN

Deletes the segment indicated by the pointer .

DELETE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DELETE ALL FILES

deletes all les .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DELETE FILE

deletes a selected le .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

sets the limits an equal amount above and below a speci ed

DELTA LIMITS

middle value , instead of setting upper and lower limits

separately . This is used in conjunction with

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER ! MIDDLE , to set limits for testing a device that is

speci ed at a particular value plus or minus an equal tolerance .

F or example , a device may be speci ed at 0 dB 3 dB . Enter

the delta limits as 3 dB and the middle value as 0 dB .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DEMODAMPLITUDE (Option 010 only) amplitude demodulation for CW time

transform measurements .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DEMOD: OFF (Option 010 only) turns time domain demodulation o .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

(Option 010 only) phase demodulation for CW TIME transform

DEMOD: PHASE

measurements .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DIRECTORY SIZE lets you specify the number of directory les to be initialized

on a disk. This is particularly useful with a hard disk, where

you may want a directory larger than the default 256 les , or

with a oppy disk you may want to reduce the directory to

allow extra space for data les . The number of directory les

must be a multiple of 8. The minimum number is 8, and there

9-10 K ey De nitions

MIDDLE VALUE or

Page 497 of the Agilent Technologies 8753D user manual — DOWN CONVERTER
Page 497 · click to enlarge
Page 497

is no practical maximum limit. Set the directory size before

initializing a disk.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DISK UNIT NUMBER

speci es the number of the disk unit in the disk drive that is to

be accessed in an external disk store or load routine . This is

used in conjunction with the HP-IB address of the disk drive ,

and the volume number , to gain access to a speci c area on a

disk. The access hierarchy is HP-IB address , disk unit number ,

disk volume number .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DISP MKRS ON off displays response and stimulus values for all markers that are

turned on. A vailable only if no marker functions are on.

DISPLA Y

provides access to a series of menus for instrument and active

channel display functions . The rst menu de nes the displayed

active channel trace in terms of the mathematical relationship

between data and trace memory . Other functions include dual

channel display (overlaid or split), display intensity , color

selection, active channel display title , and frequency blanking.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DISPLAY: DATA

displays the current measurement data for the active channel.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

leads to a series of service tests for the display .

DISPLAY TESTS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DO BOTH FWD + REV

activates both forward and reverse calibration measurements

from selected calibration menus .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DO SEQUENCE has two functions:

It shows the current sequences in memory . T o run a

sequence , press the softkey next to the desired sequence title .

When entered into a sequence , this command performs a

one-way jump to the sequence residing in the speci ed

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

sequence position (SEQUENCE 1 through 6).

DO SEQUENCE

jumps to a softkey position, not to a speci c sequence title .

Whatever sequence is in the selected softkey position will

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DO SEQUENCE command is executed. This

command prompts the operator to select a destination

sequence position.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

nishes one-port calibration (after all standards are measured)

DONE 1-PORT CAL

and turns error correction on.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DONE 2-PORT CAL nishes two-port calibration (after all standards are measured)

and turns error correction on.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DONE RESP ISOL'N CAL nishes response and isolation calibration (after all standards

are measured) and turns error correction on.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

nishes response calibration (after all standards are measured)

DONE RESPONSE

and turns error correction on.

N

DONE SEQ MODIFY terminates the sequencing edit mode .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

nishes TRL/LRM two-port calibration (after all standards are

DONE TRL/LRM

measured) and turns error correction on.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

sets the analyzer's source higher than the analyzer's receiver

DOWN CONVERTER

for making measurements in frequency o set mode .

K ey De nitions 9-11

· run when the

Page 498 of the Agilent Technologies 8753D user manual — EXT SOURCE MANUAL
Page 498 · click to enlarge
Page 498

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DUAL CH on OFF

toggles between display of both measurement channels or

the active channel only . This is used in conjunction with

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SPLIT DISP ON off in the display more menu to display both

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

channels . With

SPLIT DISP OFF the two traces are overlaid

on a single graticule .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

duplicates a sequence currently in memory into a di erent

DUPLICATE SEQUENCE

softkey position. Duplicating a sequence is straightforward.

F ollow the prompts on the analyzer screen. This command does

not a ect the original sequence .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EACH SWEEP

P ower meter calibration occurs on each sweep . Each

measurement point is measured by the power meter , which

provides the analyzer with the actual power reading. The

analyzer corrects the power level at that point. The number

of measurement/correction iterations performed on each

point is determined by the

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NUMBER OF READINGS softkey .

This measurement mode sweeps slowly , especially when the

measured power is low . Low power levels require more time for

the power meter to settle . The power meter correction table in

memory is updated after each sweep . This table can be read or

changed via HP-IB .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

displays a table of limit segments on the LCD , superimposed on

EDIT LIMIT LINE

the trace . The edit limits menu is presented so that limits can

be de ned or changed. It is not necessary for limit lines or limit

testing to be on while limits are de ned.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

presents the edit list menu. This is used in conjunction with

EDIT LIST

the edit subsweep menu to de ne or modify the frequency

sweep list. The list frequency sweep mode is selected with the

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LIST FREQ softkey described below .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

adjusts the electrical delay to balance the phase of the DUT . It

ELECTRICAL DELAY

simulates a variable length lossless transmission line , which can

be added to or removed from a receiver input to compensate

for interconnecting cables , etc . This function is similar to

the mechanical or analog \line stretchers" of other network

analyzers . Delay is annotated in units of time with secondary

labeling in distance for the current velocity factor .

N

EMIT BEEP

NNNNNNNNNNNNNNNNNNNNNNNNNNNN

causes the instrument to beep once .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

terminates the HP-GL \LB" command.

END OF LABEL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

END SWEEP HIGH PULSE

sets the TTL output on the test set interconnect to normally

high with a 10 s pulse high at the end of each sweep .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

END SWEEP LOW PULSE

sets the TTL output on the test set interconnect to normally low

with a 10 s pulse low at the end of each sweep .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ERASE TITLE

deletes the entire title .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EXECUTE TEST

runs the selected service test.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

selects the auto external source mode .

EXT SOURCE AUTO

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

selects the manual external source mode .

EXT SOURCE MANUAL

9-12 K ey De nitions

Page 499 of the Agilent Technologies 8753D user manual — FIXED MKR VALUE
Page 499 · click to enlarge
Page 499

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EXT TRIG ON POINT

is similar to the trigger on sweep , but triggers each data point in

a sweep .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EXT TRIG ON SWEEP

is used when the sweep is triggered on an externally generated

signal connected to the rear panel EXT TRIGGER input.

External trigger mode is allowed in every sweep mode .

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EXTENSION INPUT A

Use this feature to add electrical delay (in seconds) to extend

the reference plane at input A to the end of the cable . This is

used for any input measurements including S-parameters .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EXTENSION INPUT B

adds electrical delay to the input B reference plane for any B

input measurements including S-parameters .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EXTENSION PORT 1

extends the reference plane for measurements of S

, S

S

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EXTENSION PORT 2

extends the reference plane for measurements of S

, S

S

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

toggles the reference plane extension mode . When this function

is on, all extensions de ned above are enabled; when o , none

of the extensions are enabled.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EXTERNAL DISK selects an (optional) external disk drive for SA VE/RECALL.

N

EXTERNAL TESTS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

leads to a series of service tests .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

appears during sequence modi cation, when external disk

FILETITLE FILE0

is selected. FILE0 is the default name . A new name can be

entered when you save the state to disk.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FILE NAME FILE0

supplies a name for the saved istate and or data le . Brings up

the TITLE FILE MENU .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FILE UTILITIES provides access to the le utilities menu.

NNNNNNNNNNNNNNNNN

de nes the load in a calibration kit as a xed (not sliding) load.

FIXED

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FIXED MKR AUX VALUE

is used only with a polar or Smith format. It changes the

auxiliary response value of the xed marker . This is the second

part of a complex data pair , and applies to a magnitude/phase

marker , a real/imaginary marker , an R+jX marker , or a G+jB

marker . Fixed marker auxiliary response values are always

uncoupled in the two channels .

T o read absolute active marker auxiliary values following a

NNNNNNNNNNNNNNNNNNNNNNNNNN

MKR ZERO operation, the auxiliary value can be reset to zero .

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

leads to the xed marker menu, where the stimulus and

FIXED MKR POSITION

response values for a xed reference marker can be set

arbitrarily .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

changes the stimulus value of the xed marker . Fixed marker

FIXED MKR STIMULUS

stimulus values can be di erent for the two channels if the

channel markers are uncoupled using the marker mode menu.

T o read absolute active marker stimulus values following a

NNNNNNNNNNNNNNNNNNNNNNNNNN

MKR ZERO operation, the stimulus value can be reset to zero .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

changes the response value of the xed marker . In a Cartesian

FIXED MKR VALUE

format this is the y-axis value . In a polar or Smith chart format

with a magnitude/phase marker , a real/imaginary marker , an

K ey De nitions 9-13

, and · , and · EXTENSIONS on OFF

Page 500 of the Agilent Technologies 8753D user manual — FWD ISOL'N ISOL'N STD
Page 500 · click to enlarge
Page 500

R+jX marker , or a G+jB marker , this applies to the rst part of

the complex data pair . Fixed marker response values are always

uncoupled in the two channels .

T o read absolute active marker response values following a

NNNNNNNNNNNNNNNNNNNNNNNNNN

MKR ZERO operation, the response value can be reset to zero .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FLAT LINE

de nes a at limit line segment whose value is constant with

frequency or other stimulus value . This line is continuous

to the next stimulus value , but is not joined to a segment

with a di erent limit value . If a at line segment is the nal

segment it terminates at the stop stimulus . A at line segment

is indicated as FL on the table of limits .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FORM FEED puts a form feed command into the display title .

presents a menu used to select the display format for the data.

F ORMA T

V arious rectangular and polar formats are available for display

of magnitude , phase , impedance , group delay , real data, and

SWR.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FORMAT ARY on OFF speci es whether or not to store the formatted data on disk

with the instrument state .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FORMAT DISK brings up a menu for formatting a disk.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FORMAT: DOS

causes subsequent disk initialization to use the DOS disk format.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

causes subsequent disk initialization to use the LIF disk format.

FORMAT: LIF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FORMAT: LIF is the default setting.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FORMAT EXT DISK

initializes media in external drive , and formats the disk using

the selected (DOS or LIF) format.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FORMAT INT DISK

initializes media in internal drive , and formats the disk using

the selected (DOS or LIF) format.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FORMAT INT MEMORY

clears all internal save registers and associated cal data and

memory traces .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FREQ OFFS MENU

leads to the frequency o set menu.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

switches the frequency o set mode on and o .

FREQ OFFS on OFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FREQUENCY

speci es the frequency of a calibration factor or loss value in

the power meter cal loss/sensor lists .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FREQUENCY BLANK

blanks the displayed frequency notation for security purposes .

Frequency labels cannot be restored except by instrument

preset or turning the power o and then on.

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FREQUENCY: CW

sets the LO frequency to CW mode for frequency o set.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FREQUENCY: SWEEP

sets the LO frequency to sweep mode for frequency o set.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

provides access to the series of menus used to perform a

FULL 2-PORT

complete calibration for measurement of all four S-parameters

of a two-port device . This is the most accurate calibration for

measurements of two-port devices .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

measures the forward isolation of the calibration standard.

FWD ISOL'N ISOL'N STD

9-14 K ey De nitions

Page 501 of the Agilent Technologies 8753D user manual — N
Page 501 · click to enlarge
Page 501

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FWD MATCH (Label Class)

lets you enter a label for the forward match class . The label

appears during a calibration that uses this class .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FWD MATCH (Specify Class)

speci es which standards are in the forward match class in the

calibration kit.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FWD MATCH THRU is used to enter the standard numbers for the forward match

(thru) calibration. (F or default kits , this is the thru.)

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FWD TRANS (Label Class)

lets you enter a label for the forward transmission class . The

label appears during a calibration that uses this class .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FWD TRANS (Specify Class)

speci es which standards are in the forward transmission class

in the calibration kit.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FWD TRANS THRU measures the forward frequency response in a two-port

calibration.

NNNNNNNNNNNNNNNNNNNNNNNNNN

G+jB MKR displays the complex admittance values of the active marker in

rectangular form. The active marker values are displayed in

terms of conductance (in Siemens), susceptance , and equivalent

capacitance or inductance . Siemens are the international units

of admittance , and are equivalent to mhos (the inverse of

ohms). The Smith chart graticule is changed to admittance

form.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

(Option 010 only) turns gating on or o in time domain mode .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

(Option 010 only) allows you to specify the time at the center of

GATE: CENTER

the gate .

N

GATE: SPAN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

(Option 010 only) allows you to specify the gate periods .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GATE: START

(Option 010 only) allows you to specify the starting time of the

gate .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GATE: STOP

(Option 010 only) allows you to specify the stopping time of the

gate .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

(Option 010 only) leads to the gate shape menu.

GATE SHAPE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GATE SHAPE MAXIMUM

(Option 010 only) selects the widest time domain gate with the

smallest passband ripple .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GATE SHAPE MINIMUM

(Option 010 only) selects the narrowest time domain gate with

the largest passband ripple .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GATE SHAPE NORMAL

(Option 010 only) selects an intermediate time domain gate .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

(Option 010 only) selects an intermediate time domain gate .

GATE SHAPE WIDE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GET SEQ TITLES

copies the sequence titles currently in memory into the six

softkey positions .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GOSUB SEQUENCE

calls sub-routines in sequencing.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

speci es whether or not to store display graphics on disk with

the instrument state .

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GRATICULE [ ] brings up the color de nition menu. The graticule trace default

color is cyan.

K ey De nitions 9-15

Show 7 more labels / short text on this page
· G/n · giga/nano (10 · / 10 · ) · GATE on OFF · GRAPHICS on OFF
Page 502 of the Agilent Technologies 8753D user manual — IF LIMIT TEST PASS
Page 502 · click to enlarge
Page 502

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GRATICULE TEXT

selects the graticule and a portion of softkey text (where there

is a choice of a feature being on or o ) for color modi cation.

F or example:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FREQUENCY BLANK on OFF .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

HARMONIC MEAS

(Option 002 only) leads to the harmonics menu. This feature

phase locks to the 2nd or 3rd harmonic of the fundamental

signal. Measured harmonics cannot exceed the frequency range

of the analyzer receiver .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

HARMONIC OFF

(Option 002 only) turns o the harmonic measurement mode .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

(Option 002 only) selects measurement of the second harmonic .

HARMONIC SECOND

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

HARMONIC THIRD

(Option 002 only) selects measurement of the third harmonic .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

provides an on-line quick reference guide to using the adapter

HELP ADAPT REMOVAL

removal technique .

NNNNNNNNNNNNNN

freezes the data trace on the display , and the analyzer stops

HOLD

sweeping and taking data. The notation \Hld" is displayed at

the left of the graticule . If the * indicator is on at the left side

of the display , trigger a new sweep with

.

SINGLE

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

toggles the HP-IB diagnostic feature (debug mode). This mode

HP-IB DIAG on off

should only be used the rst time a program is written: if a

program has already been debugged, it is unnecessary .

When diagnostics are on, the analyzer scrolls a history of

incoming HP-IB commands across the display in the title line .

Nonprintable characters are represented as . If a syntax error

is received, the commands halt and a pointer ^ indicates the

misunderstood character . T o clear a syntax error , refer to the

\HP-IB Programming Reference" and \HP-IB Programming

Examples" chapters in the HP 8753D Network Analyzer

Programmer's Guide .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

is used to select the bandwidth value for IF bandwidth

IF BW [ ]

reduction. Allowed values (in Hz) are 3700, 3000, 1000, 300,

100, 30, and 10. Any other value will default to the closest

allowed value . A narrow bandwidth slows the sweep speed but

provides better signal-to-noise ratio . The selected bandwidth

value is shown in brackets in the softkey label.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

jumps to one of the six sequence positions (SEQUENCE 1

IF LIMIT TEST FAIL

through 6) if the limit test fails . This command executes any

sequence residing in the selected position. Sequences may

jump to themselves as well as to any of the other sequences in

memory . When this softkey is pressed, the analyzer presents

a softkey menu showing the six sequence positions and the

titles of the sequences located in them. Choose the destination

sequence to be called if the limit test fails .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

jumps to one of the six sequence positions (SEQUENCE 1

IF LIMIT TEST PASS

through 6) if the limit test passes . This command executes any

sequence residing in the selected position. Sequences may

jump to themselves as well as to any of the other sequences in

memory . When this softkey is pressed, the analyzer presents a

softkey menu showing the six sequence positions , and the titles

9-16 K ey De nitions

Page 503 of the Agilent Technologies 8753D user manual — ISOL'N STD
Page 503 · click to enlarge
Page 503

of the sequences located in them. Choose the sequence to be

called if the limit test passes (destination sequence).

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

IF LOOP COUNTER = 0

prompts the user to select a destination sequence position

(SEQUENCE 1 through 6). When the value of the loop counter

reaches zero , the sequence in the speci ed position will run.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

IF LOOP < > COUNTER 0

prompts the user to select a destination sequence position

(SEQUENCE 1 through 6). When the value of the loop counter

is no longer zero , the sequence in the speci ed position will run.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

IMAGINARY

displays only the imaginary (reactive) portion of the measured

data on a Cartesian format. This format is similar to the real

format except that reactance data is displayed on the trace

instead of impedance data.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INCR LOOP COUNTER

increments the value of the loop counter by 1.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

initializes the disk unit number and volume number selected in

INIT DISK? YES

the HP-IB menu, then returns to the disk menu. If more than

one hard disk volume is to be initialized, each volume must be

selected and initialized individually .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INITIALIZE DISK

leads to the initialize menu. Before data can be stored on

a disk, the disk must be initialized. If you attempt to store

without initializing the disk, the message \CA UTION: DISK

MEDIUM NOT INITIALIZED" is displayed. The disk format can

be selected to be either logical interchange format (LIF), or

DOS.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INPUT PORTS accesses a menu that allows you to measure the R, A, and B

channels .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INSTRUMENT MODE presents the instrument mode menu. This provides access to

the primary modes of operation (analyzer modes).

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INTENSITY

sets the LCD intensity as a percent of the brightest setting. The

factory-set default value is stored in non-volatile memory .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INTERNAL TESTS

leads to a series of service tests .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

selects the analyzer internal disk for the storage device .

INTERNAL DISK

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INTERNAL MEMORY

selects internal non-volatile memory as the storage medium for

subsequent save and recall activity .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

turns interpolated error correction on or o . The interpolated

error correction feature allows the operator to calibrate the

system, then select a subset of the frequency range or a

di erent number of points . Interpolated error correction

functions in linear frequency , power sweep and CW time

modes . When using the analyzer in linear sweep , it is

recommended that the original calibration be performed with at

least 67 points per 1 GHz of frequency span.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

leads to the isolation menu.

ISOLATION

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ISOLATION DONE

returns to the two-port cal menu.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

measures the isolation of the device connected to the test port.

ISOL'N STD

K ey De nitions 9-17

INTERPOL on OFF

Page 504 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 504 · click to enlarge
Page 504

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ISTATE CONTENTS

describes the selected instrument state le (disk only)

translating the various lename pre xes into more descriptive

detail.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

KIT DONE (MODIFIED)

terminates the cal kit modi cation process , after all standards

are de ned and all classes are speci ed. Be sure to save the kit

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SAVE USER KIT softkey , if it is to be used later .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LABEL CLASS

leads to the label class menu, to give the class a meaningful

label for future reference during calibration.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LABEL CLASS DONE

nishes the label class function and returns to the modify cal

kit menu.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LABEL KIT leads to a menu for constructing a label for the user-modi ed

cal kit. If a label is supplied, it will appear as one of the ve

softkey choices in the select cal kit menu. The approach is

similar to de ning a display title , except that the kit label is

limited to ten characters .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LABEL STD The function is similar to de ning a display title , except that

the label is limited to ten characters .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LEFT LOWER draws a quarter-page plot in the lower left quadrant of the

page .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

draws a quarter-page plot in the upper left quadrant of the

LEFT UPPER

page .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

leads to the o set limits menu, which is used to o set the

LIMIT LINE OFFSETS

complete limit set in either stimulus or amplitude value .

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

turns limit lines on or o . T o de ne limits , use the

LIMIT LINE on OFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EDIT LIMIT LINE softkey described below . If limits have

been de ned and limit lines are turned on, the limit lines are

displayed on the LCD for visual comparison of the measured

data in all Cartesian formats .

If limit lines are on, they are plotted with the data on a plot,

and saved in memory with an instrument state . In a listing of

values from the copy menu with limit lines on, the upper limit

and lower limit are listed together with the pass or fail margin,

as long as other listed data allows sucient space .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LIMIT MENU

leads to a series of menus used to de ne limits or speci cations

with which to compare a test device . Refer to Limit Lines and

Limit T esting.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LIMIT TEST on OFF

turns limit testing on or o . When limit testing is on, the data

is compared with the de ned limits at each measured point.

Limit tests occur at the end of each sweep , whenever the data

is updated, when formatted data is changed, and when limit

testing is rst turned on.

Limit testing is available for both magnitude and phase values

in Cartesian formats . In polar and Smith chart formats , the

value tested depends on the marker mode and is the magnitude

or the rst value in a complex pair . The message \NO LIMIT

9-18 K ey De nitions

Show 6 more labels / short text on this page
· k/m · kilo/milli (10 · / 10 · ) · with the
Page 505 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 505 · click to enlarge
Page 505

LINES DISPLA YED" is displayed in polar and Smith chart

formats if limit lines are turned on.

Five indications of pass or fail status are provided when limit

testing is on. A P ASS or F AIL message is displayed at the right

of the LCD . The trace vector leading to any measured point that

is out of limits is set to red at the end of every limit test, both

on a displayed plot and a hard copy plot. The limit fail beeper

sounds if it is turned on. In a listing of values using the copy

menu, an asterisk * is shown next to any measured point that is

out of limits . A bit is set in the HP-IB status byte .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LIMIT TEST RESULT

puts the result of a limit test into the display title .

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

leads to the limit type menu, where one of three segment types

LIMIT TYPE

can be selected.

NNNNNNNNNNNNNNNNNNNNNNNNNN

LIN FREQ activates a linear frequency sweep displayed on a standard

graticule with ten equal horizontal divisions . This is the default

preset sweep type .

NNNNNNNNNNNNNNNNNNNNNNN

LIN MAG

displays the linear magnitude format. This is a Cartesian format

used for unitless measurements such as re ection coecient

magnitude or transmission coecient magnitude , and for

linear measurement units . It is used for display of conversion

parameters and time domain transform data.

NNNNNNNNNNNNNNNNNNNNNNN

LIN MKR

displays a readout of the linear magnitude and the phase of the

active marker . Marker magnitude values are expressed in units;

phase is expressed in degrees .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LINE/MATCH

provides access to the Line/Match Menu for TRL/LRM

calibration.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LINE TYPE DATA selects the line type for the data trace plot. The default line

type is 7, which is a solid unbroken line .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LINE TYPE MEMORY

selects the line type for the memory trace plot. The default line

type is 7.

NNNNNNNNNNNNNN

LIST

provides a tabular listing of all the measured data points and

their current values , together with limit information if it is

turned on. At the same time , the screen menu is presented, to

enable hard copy listings and access new pages of the table . 30

lines of data are listed on each page , and the number of pages is

determined by the number of measurement points speci ed in

the stimulus menu.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LIST FREQ provides a user-de nable arbitrary frequency list mode . This

list is de ned and modi ed using the edit list menu and the

edit subsweep menu. Up to 30 frequency subsweeps (called

\segments") of several di erent types can be speci ed, for

a maximum total of 1632 points . One list is common to both

channels . Once a frequency list has been de ned and a

measurement calibration performed on the full frequency list,

one or all of the frequency segments can be measured and

displayed without loss of calibration.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LIST VALUES

provides a tabular listing of all the measured data points and

their current values , together with limit information if it is

switched on. Thirty lines of data are listed on each page ,

K ey De nitions 9-19

Page 506 of the Agilent Technologies 8753D user manual — LOOP COUNTER
Page 506 · click to enlarge
Page 506

and the number of pages is determined by the number of

measurement points speci ed in the stimulus menu.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LN/MATCH 1

measures the TRL/LRM line or match standard for PORT 1.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LN/MATCH 2

measures the TRL/LRM line or match standard for PORT 2.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

turns the LO control mode on and o for frequency o set.

LO CONTROL on OFF

N

LO MENU

NNNNNNNNNNNNNNNNNNNNNN

leads to the LO menu. Allows you to con gure the external

source for frequency o set.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LO SOURCE ADDRESS

shows the HP-IB address of the LO source .

NNNNNNNNNNNNNN

LOAD

de nes the standard type as a load (termination). Loads

are assigned a terminal impedance equal to the system

characteristic impedance Z0, but delay and loss o sets may still

be added. If the load impedance is not Z0, use the arbitrary

impedance standard de nition.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOAD NO OFFSET

initiates measurement of a calibration standard load without

o set.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOAD OFFSET initiates measurement of a calibration standard load with o set.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

presents the load sequence from disk menu. Select the desired

LOAD SEQ FROM DISK

sequence and the analyzer will load it from disk.

This key is used to return the analyzer to local (front panel)

LOCAL

operation from remote (computer controlled) operation. This

key will also abort a test sequence or hardcopy print/plot. In

this local mode , with a controller still connected on HP-IB , the

analyzer can be operated manually (locally) from the front

panel. This is the only front panel key that is not disabled

when the analyzer is remotely controlled over HP-IB by a

computer . The exception to this is when local lockout is in

e ectthis is a remote command that disables the

LOCAL

making it dicult to interfere with the analyzer while it is

under computer control.

NNNNNNNNNNNNNNNNNNNNNNNNNN

activates a logarithmic frequency sweep mode . The source is

LOG FREQ

stepped in logarithmic increments and the data is displayed on

a logarithmic graticule . This is slower than a continuous sweep

with the same number of points , and the entered sweep time

may therefore be changed automatically . F or frequency spans

of less than two octaves , the sweep type automatically reverts

to linear sweep .

NNNNNNNNNNNNNNNNNNNNNNN

displays the log magnitude format. This is the standard

LOG MAG

Cartesian format used to display magnitude-only measurements

of insertion loss , return loss , or absolute power in dB versus

frequency .

NNNNNNNNNNNNNNNNNNNNNNN

displays the logarithmic magnitude value and the phase of the

LOG MKR

active marker in P olar or Smith chart format. Magnitude values

are expressed in dB and phase in degrees . This is useful as a

fast method of obtaining a reading of the log magnitude value

without changing to log magnitude format.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

displays the current value of the loop counter and allows you

LOOP COUNTER

to change the value of the loop counter . Enter any number

9-20 K ey De nitions

· key ,

Page 507 of the Agilent Technologies 8753D user manual — MARKER
Page 507 · click to enlarge
Page 507

from 0 to 32767 and terminate with the

key . The default

value of the counter is zero . This command should be placed in

a sequence that is separate from the measurement sequence .

F or this reasonthe measurement sequence containing a loop

decision command must call itself in order to function. The

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOOP COUNTER command must be in a separate sequence or the

counter value would always be reset to the initial value .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOOP COUNTER

inserts the string \[LOOP]" into the lename .

N

NNNNNNNNNNNNN

accepts a power loss value for a segment in the power meter

LOSS

cal power loss list. This value , for example , could be the

di erence (in dB) between the coupled arm and through arm of

a directional coupler .

N

LOSS/SENSR LISTS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

presents the power loss/sensor lists menu. This menu performs

two functions:

Corrects coupled-arm power loss when a directional coupler is

used to sample the RF output.

Allows calibration factor data to be entered for one or two

power sensors .

Each function provides up to 12 separate frequency points ,

called segments , at which the user may enter a di erent power

loss or calibration factor . The instrument interpolates between

the selected points . Two power sensor lists are provided

because no single power sensor can cover the frequency range

possible with an HP 8753D .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOW PASS IMPULSE

(Option 010 only) sets the transform to low pass impulse mode ,

which simulates the time domain response to an impulse input.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOW PASS STEP

(Option 010 only) sets the transform to low pass step mode ,

which simulates the time domain response to a step input.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOWER LIMIT

sets the lower limit value for the start of the segment in a

limit line list. If an upper limit is speci ed, a lower limit must

also be de ned. If no lower limit is required for a particular

measurement, force the lower limit value out of range (for

example 500 dB).

M/

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MANUAL TRG ON POINT

waits for a manual trigger for each point. Subsequent pressing

of this softkey triggers each measurement. The annotation

\man" will appear at the left side of the display when the

instrument is waiting for the trigger to occur . This feature

is useful in a test sequence when an external device or

instrument requires changes at each point.

MARKER

displays an active marker on the screen and provides access to

a series of menus to control from one to ve display markers

for each channel. Markers provide numerical readout of

measured values at any point of the trace .

The menus accessed from the

MARKER

key provide several

basic marker operations . These include special marker modes

for di erent display formats , and a marker delta mode that

K ey De nitions 9-21

Show 6 more labels / short text on this page
(Sequence Filenaming) · · mega/micro (10 · / 10 · ) ·
Page 508 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 508 · click to enlarge
Page 508

displays marker values relative to a speci ed value or another

marker .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER ! AMP. OFS. uses the active marker to set the amplitude o set for the limit

lines . Move the marker to the desired middle value of the limits

and press this softkey . The limits are then moved so that they

are centered an equal amount above and below the marker at

that stimulus value .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER ! CENTER changes the stimulus center value to the stimulus value of the

active marker , and centers the new span about that value .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER ! CW

sets the CW frequency of the analyzer to the frequency

of the active marker . This feature is intended for use in

automated compression measurements . T est sequences allow

the instrument to automatically nd a maximum or minimum

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

point on a response trace . The

MARKER ! CW command sets

the instrument to the CW frequency of the active marker .

When power sweep in engaged, the CW frequency will already

be selected.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER ! DELAY adjusts the electrical delay to balance the phase of the DUT .

This is performed automatically , regardless of the format and

the measurement being made . Enough line length is added to or

subtracted from the receiver input to compensate for the phase

slope at the active marker position. This e ectively attens

the phase trace around the active marker , and can be used

to measure electrical length or deviation from linear phase .

A dditional electrical delay adjustments are required on DUT s

without constant group delay over the measured frequency

span. Since this feature adds phase to a variation in phase

versus frequency , it is applicable only for ratioed inputs .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER ! MIDDLE sets the midpoint for

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DELTA LIMITS using the active marker

to set the middle amplitude value of a limit segment. Move the

marker to the desired value or device speci cation, and press

this key to make that value the midpoint of the delta limits .

The limits are automatically set an equal amount above and

below the marker .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER ! REFERENCE makes the reference value equal to the active marker's

response value , without changing the reference position. In a

polar or Smith chart format, the full scale value at the outer

circle is changed to the active marker response value . This

softkey also appears in the scale reference menu.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

changes the start and stop values of the stimulus span to the

MARKER ! SPAN

values of the active marker and the delta reference marker . If

there is no reference marker , the message \NO MARKER DEL T A

SP AN NOT SET" is displayed.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

changes the stimulus start value to the stimulus value of the

MARKER ! START

active marker .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER ! STIMULUS sets the starting stimulus value of a limit line segment using the

active marker . Move the marker to the desired starting stimulus

value before pressing this key , and the marker stimulus value is

entered as the segment start value .

9-22 K ey De nitions

Page 509 of the Agilent Technologies 8753D user manual — MAXIMUM FREQUENCY
Page 509 · click to enlarge
Page 509

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER ! STOP

changes the stimulus stop value to the stimulus value of the

active marker .

NNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER 1 turns on marker 1 and makes it the active marker . The active

marker appears on the display as r . The active marker

stimulus value is displayed in the active entry area, together

with the marker number . If there is a marker turned on, and no

other function is active , the stimulus value of the active marker

can be controlled with the knob , the step keys , or the numeric

keypad. The marker response and stimulus values are displayed

in the upper right-hand corner of the screen.

NNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER 2

turns on marker 2 and makes it the active marker . If another

marker is present, that marker becomes inactive and is

represented on the display as .

NNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER 3

turns on marker 3 and makes it the active marker .

NNNNNNNNNNNNNNNNNNNNNNNNNN

turns on marker 4 and makes it the active marker .

MARKER 4

NNNNNNNNNNNNNNNNNNNNNNNNNN

turns on marker 5 and makes it the active marker .

MARKER 5

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

turns o all the markers and the delta reference marker , as well

as the tracking and bandwidth functions that are accessed with

NNNNNNNNNNNNNNNNNNNNNNNNNN

MKR FCTN key .

key activates a marker if one is not already active , and provides

MARKER F CTN

access to additional marker functions . These can be used

to quickly change the measurement parameters , to search

the trace for speci ed information, and to analyze the trace

statistically .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

provides access to the marker mode menu, where several

MARKER MODE MENU

marker modes can be selected including special markers for

polar and Smith chart formats .

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

located under the

key , interpolates between measured

MARKERS: CONTINUOUS

MARKER

points to allow the markers to be placed at any point on the

trace . Displayed marker values are also interpolated. This is the

default marker mode .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKERSCOUPLED couples the marker stimulus values for the two display

channels . Even if the stimulus is uncoupled and two sets of

stimulus values are shown, the markers track the same stimulus

values on each channel as long as they are within the displayed

stimulus range .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

places markers only on measured trace points determined by

MARKERS: DISCRETE

the stimulus settings .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKERS: UNCOUPLED

allows the marker stimulus values to be controlled

independently on each channel.

NNNNNNNNNNN

MAX

moves the active marker to the maximum point on the trace .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

is used to de ne the highest frequency at which a calibration

MAXIMUM FREQUENCY

kit standard can be used during measurement calibration. In

waveguide , this is normally the upper cuto frequency of the

standard.

K ey De nitions 9-23

MARKER all OFF · the ·

Page 510 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 510 · click to enlarge
Page 510

key provides access to a series of softkey menus for selecting

MEAS

the parameters or inputs to be measured.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

aborts the sweep in progress , then restarts the measurement.

MEASURE RESTART

This can be used to update a measurement following an

adjustment of the device under test. When a full two-port

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

calibration is in use , the

MEASURE RESTART key will initiate

another update of both forward and reverse S-parameter data.

This softkey will also override the test set hold mode , which

inhibits continuous switching of either the test port transfer

switch or step attenuator . The measurement con gurations

which cause this are described in T est Set Attenuator , T est

P ort Transfer Switch, and Doubler Switch Protection, at the

beginning of this section. This softkey will override the test set

hold mode for one measurement.

If the analyzer is taking a number of groups (see Trigger

Menu), the sweep counter is reset at 1. If averaging is on,

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MEASURE RESTART resets the sweep-to-sweep averaging and is

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

e ectively the same as

AVERAGING RESTART . If the sweep

NNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MEASURE RESTART executes a single

sweep .

NNNNNNNNNNNNNNNNNNNN

displays the trace memory for the active channel. This is the

MEMORY

only memory display mode where the smoothing and gating of

the memory trace can be changed. If no data has been stored in

memory for this channel, a warning message is displayed.

provides access to a series of menus which are used to de ne

MENU

and control all stimulus functions other than start, stop , center ,

and span. When the

key is pressed, the stimulus menu is

MENU

displayed.

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MIDDLE VALUE sets the midpoint for

DELTA LIMITS . It uses the entry controls

to set a speci ed amplitude value vertically centered between

the limits .

NNNNNNNNNNN

MIN moves the active marker to the minimum point on the trace .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MINIMUM FREQUENCY

is used to de ne the lowest frequency at which a calibration

kit standard can be used during measurement calibration. In

waveguide , this must be the lower cuto frequency of the

standard, so that the analyzer can calculate dispersive e ects

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

OFFSET DELAY ).

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

leads to the marker search menu, which is used to search the

MKR SEARCH [ ]

trace for a particular value or bandwidth.

NNNNNNNNNNNNNNNNNNNNNNNNNN

puts a xed reference marker at the present active marker

MKR ZERO

position, and makes the xed marker stimulus and response

values at that position equal to zero . All subsequent stimulus

and response values of the active marker are then read out

relative to the xed marker . The xed marker is shown on the

display as a small triangle (delta), smaller than the inactive

NNNNNNNNNNNNNNNNNNNNNNNNNN

marker triangles . The softkey label changes from

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MKR ZERO REF = and the notation \REF = " is displayed

at the top right corner of the graticule . Marker zero is canceled

9-24 K ey De nitions

Show 6 more labels / short text on this page
· trigger is in · HOLD mode , · · correctly (see · MKR ZERO to
Page 511 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNN
Page 511 · click to enlarge
Page 511

by turning delta mode o in the delta marker menu or turning

NNNNNNNNNNNNNNNNNNNNNNN

all the markers o with the

ALL OFF softkey .

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MODIFY [ ]

leads to the modify cal kit menu, where a default cal kit can be

user-modi ed.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MODIFY COLORS present a menu for color modi cation of display elements .

Refer to A djusting Color for information on modifying display

elements .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NETWORK ANALYZER sets the analyzer to network analyzer mode .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NEW SEQ/MODIFY SEQ

activates the sequence edit mode and presents the new/modify

sequence menu with a list of sequences that can be created or

modi ed.

NNNNNNNNNNNNNNNNNNNNNNN

NEWLINE puts a new line command into the display title .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NEXT PAGE

steps forward through a tabular list of data page-by-page .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

triggers a user-speci ed number of sweeps , and returns to the

NUMBER OF GROUPS

hold mode . This function can be used to override the test set

hold mode , which protects the electro-mechanical transfer

switch and attenuator against continuous switching. This is

explained fully in the T est Set Attenuator description in the

\Application and Operation Concepts" chapter , in this manual.

If averaging is on, the number of groups should be at least

equal to the averaging factor selected to allow measurement of

a fully averaged trace . Entering a number of groups resets the

averaging counter to 1.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NUMBER OF POINTS

is used to select the number of data points per sweep to be

measured and displayed. Using fewer points allows a faster

sweep time but the displayed trace shows less horizontal detail.

Using more points gives greater data density and improved

trace resolution, but slows the sweep and requires more

memory for error correction or saving instrument states .

The possible values that can be entered for number of points

are 3, 11, 26, 51, 101, 201, 401,801, and 1601. The number

of points can be di erent for the two channels if the stimulus

values are uncoupled.

In list frequency sweep , the number of points displayed is the

total number of frequency points for the de ned list (see Sweep

Type Menu).

N

NUMBER OF READINGS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

determines the number of measurement/correction iterations

performed on each point in a power meter calibration. This

feature helps eliminate residual power errors after the initial

correction. The amount of residual error is directly proportional

to the magnitude of the initial correction. The user should

initially set the source power so that it is approximately

correct at the device under test. If power uncertainty at the

device under test is expected to be greater than a few dB , it is

recommended that the number of readings be greater than 1.

NNNNNNNNNNNNNNNNNNN

N

OFFSET

selects the calibration standard load as being o set.

K ey De nitions 9-25

Page 512 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 512 · click to enlarge
Page 512

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

OFFSET DELAY

is used to specify the one-way electrical delay from the

measurement (reference) plane to the standard, in seconds (s).

(In a transmission standard, o set delay is the delay from plane

to plane .) Delay can be calculated from the precise physical

length of the o set, the permittivity constant of the medium,

and the speed of light.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

OFFSET LOADS DONE

completes the selection in the O set Load Menu.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

is used to specify energy loss , due to skin e ect, along a

OFFSET LOSS

one-way length of coax o set. The value of loss is entered as

ohms/nanosecond (or Giga ohms/second) at 1 GHz. (Such losses

are negligible in waveguide , so enter 0 as the loss o set.)

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

is used to specify the characteristic impedance of the coax

OFFSET Z0

o set. (NoteThis is not the impedance of the standard itself.)

(F or waveguide , the o set impedance should always be assigned

a value equal to the system Z0.)

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

OMIT ISOLATION

is used to omit the isolation portion of the calibration.

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

leads to the series of menus used to perform a high-accuracy

ONE-PATH 2-PORT

two-port calibration without an S-parameter test set.

This calibration procedure e ectively removes directivity ,

source match, load match, isolation, re ection tracking, and

transmission tracking errors in one direction only . Isolation

correction can be omitted for measurements of devices with

limited dynamic range . (The device under test must be

manually reversed between sweeps to accomplish measurement

of both input and output responses .) The required standards are

a short, an open, a thru, and an impedance-matched load.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

This mode does not measure each sweep , but corrects each

ONE SWEEP

point with the data currently in the power meter correction

table .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

OP PARMS (MKRS etc) provides a tabular listing on the analyzer display of the key

parameters for both channels . The screen menu is presented to

allow hard copy listings and access new pages of the table . F our

pages of information are supplied. These pages list operating

parameters , marker parameters , and system parameters that

relate to control of peripheral devices rather than selection of

measurement parameters .

NNNNNNNNNNNNNN

de nes the standard type as an open, used for calibrating

OPEN

re ection measurements . Opens are assigned a terminal

impedance of in nite ohms , but delay and loss o sets may still

be added. Pressing this key also brings up a menu for de ning

the open, including its capacitance .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

gets data from an HP-IB device set to the address at which the

P MTR/HPIB TO TITLE

analyzer expects to nd a power meter . The data is stored in a

title string. The analyzer must be in system controller or pass

control mode .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PARALL IN BIT NUMBER while creating a sequence , this softkey will insert a command

that selects the single bit (0 to 4) that a sequence will be

looking for from the GPIO bus .

9-26 K ey De nitions

Page 513 of the Agilent Technologies 8753D user manual — PHASE
Page 513 · click to enlarge
Page 513

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PARALL IN IF BIT H

while creating a sequence , this softkey inserts a command to

jump to another sequence if the single input selected is in a

high state .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PARALL IN IF BIT L

while creating a sequence , this softkey inserts a command to

jump to another sequence if the single input selected is in a low

state .

NNNNNNNNNNNNNNNNNNNNNNNNNN

PARALLEL sets the printer or plotter port to parallel.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

toggles the parallel output port between the copy and GPIO

PARALLEL [COPY/GPIO]

output modes .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PARALLEL OUT ALL

allows you to input a number (0 to 255) in base 10, and outputs

it to the bus as binary , when the parallel port is in GPIO mode .

NNNNNNNNNNNNNNNNN

PAUSE

pauses the sequence so the operator can perform a needed task,

such as changing the DUT , changing the calibration standard, or

other similar task. Press

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CONTINUE SEQUENCE when ready .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PAUSE TO SELECT when editing a sequence ,

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PAUSE TO SELECT appears when you

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DO SEQUENCE . When placed in a sequence , it presents

the menu of up to 6 available sequences (softkeys containing

non-empty sequences). The message \CHOOSE ONE OF THESE

SEQUENCES" is displayed and the present sequence is stopped.

If the operator selects one of the sequences , that sequence is

executed. Any other key can be used to exit this mode . This

function is not executed if used during modify mode and

does nothing when operated manually . This softkey is not

visible on the display , and the function is not available , unless

programmed into analyzer memory .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PEN NUM DATA

selects the number of the pen to plot the data trace . The

default pen for channel 1 is pen number 2, and for channel 2 is

pen number 3.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PEN NUM GRATICULE

selects the number of the pen to plot the graticule . The default

pen for channel 1 is pen number 1, and for channel 2 is pen

number 1.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PEN NUM MARKER selects the number of the pen to plot both the markers and the

marker values . The default pen for channel 1 is pen number 7,

and for channel 2 is pen number 7.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PEN NUM MEMORY

selects the number of the pen to plot the memory trace . The

default pen for channel 1 is pen number 5, and for channel 2 is

pen number 6.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PEN NUM TEXT

selects the number of the pen to plot the text. The default pen

for channel 1 is pen number 7, and for channel 2 is pen

number 7.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PHASE OFFSET adds or subtracts a phase o set that is constant with frequency

(rather than linear). This is independent of

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER ! DELAY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ELECTRICAL DELAY .

N

NNNNNNNNNNNNNNNN

(Option 010 only) displays a Cartesian format of the phase

PHASE

portion of the data, measured in degrees . This format displays

the phase shift versus frequency .

K ey De nitions 9-27

press · and

Page 514 of the Agilent Technologies 8753D user manual — POLAR
Page 514 · click to enlarge
Page 514

NNNNNNNNNNNNNN

PLOT

makes a hard copy plot of one page of the tabular listing on

the display , using a compatible HP plotter connected to the

analyzer through HP-IB . This method is appropriate when speed

of output is not a critical factor .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOT DATA ON off

speci es whether the data trace is to be drawn (on) or not

drawn (o ) on the plot.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOT GRAT ON off

speci es whether the graticule and the reference line are

to be drawn (on) or not drawn (o ) on the plot. Turning

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOT GRAT ON and all other elements o is a convenient

way to make preplotted grid forms . However , when data is to

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

be plotted on a preplotted form,

PLOT GRAT OFF should be

selected.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOT MEM ON off speci es whether the memory trace is to be drawn (on) or not

drawn (o ) on the plot. Memory can only be plotted if it is

displayed (refer to \Display Menu" in Chapter 6).

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOT MKR ON off

speci es whether the markers and marker values are to be

drawn (on) or not drawn (o ) on the plot.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOT NAME PLOTFILE

supplies a name for the plot le generated by a PLOT to disk.

Brings up the TITLE FILE MENU .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOT SPEED [ ]

toggles between fast and slow speeds .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

selects plotting of all displayed text except the marker values ,

PLOT TEXT ON off

softkey labels , and display listings such as the frequency list

table or limit table . (Softkey labels can be plotted under the

control of an external controller . Refer to the Introductory

Programming Guide .)

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOTTER BAUD RATE

sets the serial port data transmission speed for plots .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

sends a page eject command to the plotter .

PLOTTER FORM FEED

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

sets the HP-IB address the analyzer will use to communicate

PLOTTER PORT

with the plotter .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLTR PORT: DISK

directs plots to the selected disk (internal or external).

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

directs plots to the HP-IB port and sets the HP-IB address the

PLTR PORT: HPIB

analyzer will use to communicate with the plotter .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

con gures the analyzer for a plotter that has a parallel

PLTR PORT: PARALLEL

(centronics) interface .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

con gures the analyzer for a plotter that has a serial (RS-232)

PLTR PORT: SERIAL

interface .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLTR TYPE [PLOTTER] selects a pen plotter such as the HP 7440A, HP 7470A,

HP 7475A, or HP 7550B as the plotter type .

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

selects a PCL5 compatible printer , which supports HP-GL/2,

PLTR TYPE [HPGL PRT]

such as the LaserJet III or LaserJet 4 for a monochrome plotter

type , or the DeskJet 1200C for a color plotter type .

NNNNNNNNNNNNNNNNN

displays a polar format. Each point on the polar format

POLAR

corresponds to a particular value of both magnitude and phase .

Quantities are read vectorallythe magnitude at any point is

determined by its displacement from the center (which has zero

9-28 K ey De nitions

Page 515 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNN
Page 515 · click to enlarge
Page 515

value), and the phase by the angle counterclockwise from the

positive x-axis . Magnitude is scaled in a linear fashion, with the

value of the outer circle usually set to a ratio value of 1. Since

there is no frequency axis , frequency information is read from

the markers .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

POLAR MKR MENU leads to a menu of special markers for use with a polar format.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PORT EXTENSIONS

goes to the reference plane menu, which is used to extend the

apparent location of the measurement reference plane or input.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PORT PWR [COUPLED]

is used to set the same power levels at each port.

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

allows you to set di erent power levels at each port.

PORT PWR [UNCOUPLED]

NNNNNNNNNNNNNNNNN

POWER

makes power level the active function and sets the RF

output power level of the analyzer's internal source . The

analyzer will detect an input power overload at any of the

three receiver inputs , and automatically reduce the output

power of the source to 85 dBm. This is indicated with the

message \O VERLO AD ON INPUT (R, A, B)." In addition,

the annotation \P# " appears at the left side of the display .

When this occurs , set the power to a lower level, and toggle

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SOURCE PWR on OFF . If power meter cal is on, cal power is

the active entry .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

POWER: FIXED

sets the external LO xed power .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

sets the external LO power sweep .

POWER: SWEEP

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

brings up the segment modify menu and segment edit (power

POWER LOSS

loss) menu explained in the following pages . This softkey is

intended for use when the power output is being sampled

by a directional coupler or power splitter . In the case of the

directional coupler , enter the power loss caused by the coupled

arm. Refer to P ower Loss F eature on a previous page .

This feature may be used to compensate for attenuation

non-linearities in either a directional coupler or a power splitter .

Up to 12 segments may be entered, each with a di erent

frequency and power loss value .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNN

POWER MTR toggles between

438A/437 . These power meters are

HP-IB compatible with the analyzer . The model number in the

softkey label must match the power meter to be used.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

turns on a power sweep mode that is used to characterize

POWER SWEEP

power-sensitive circuits . In this mode , power is swept at a

single frequency , from a start power value to a stop power

value , selected using the

ST ART

STOP

keys and the entry

block. This feature is convenient for such measurements as

gain compression or A GC (automatic gain control) slope . T o

NNNNNNNNNNNNNNNNNNNNNNN

set the frequency of the power sweep , use

CW FREQ in the

stimulus menu. Refer to the User's Guide for an example of a

gain compression measurement.

mode .

K ey De nitions 9-29

436A or · and

Page 516 of the Agilent Technologies 8753D user manual — PRNTR TYPE [THINKJET]
Page 516 · click to enlarge
Page 516

In power sweep , the entered sweep time may be automatically

changed if it is less than the minimum required for the current

con guration (number of points , IF bandwidth, averaging, etc .).

PRESET

presents a menu to select a factory or user de ned preset state .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRESET: FACTORY

is used to select the preset conditions de ned by the factory .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

is used to select a preset condition de ned by the user . This

PRESET: USER

is done by saving a state in a register under

SA VE/RECALL

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

and naming the register UPRESET . When

PRESETUSER is

PRESET

key will bring up the state of the

UPRESET register .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PREVIOUS PAGE

steps backward through a tabular list of data page-by-page .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

when displaying list values , prints the entire list in color . When

PRINT ALL COLOR

displaying operating parameters , prints all but the last page in

color . The data is sent to the printer as ASCII text rather than

as raster graphics , which causes the printout to be faster .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRINT ALL MONOCHROME

when displaying list values , prints the entire list in

monochrome . When displaying operating parameters , prints

all but the last page in monochrome . The data is sent to the

printer as ASCII text rather than as raster graphics , which

causes the printout to be faster .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRINT: COLOR

sets the print command to default to a color printer . The

printer output is always in the analyzer default color values .

This command does not work with a black and white printer .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRINT COLOR

prints the displayed measurement results in color .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

is used to select the print colors menu.

PRINT COLORS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRINT: MONOCHROME

sets the print command to default to a black and white printer .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

prints the displayed measurement results in black and white .

PRINT MONOCHROME

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

prints any sequence currently in memory to a compatible

PRINT SEQUENCE

printer .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRINTER BAUD RATE

sets the serial port data transmission speed for prints .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

sends a conditional form feed to the printer .

PRINTER FORM FEED

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRINTER PORT

sets the HP-IB address the analyzer will use to communicate

with the printer .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRNTR TYPE [DESKJET]

sets the printer type to the DeskJet series .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

sets the printer type to Epson compatible printers , which

PRNTR TYPE [EPSON-P2]

support the Epson ESC/P2 printer control language .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

sets the printer type to the LaserJet series .

PRNTR TYPE [LASERJET]

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRNTR TYPE [PAINTJET]

sets the printer type to the P aintJet.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

sets the printer type to the ThinkJet or QuietJet.

PRNTR TYPE [THINKJET]

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PWR LOSS on OFF

turns on or o power loss correction. P ower loss correction

should be used when the power output is measured by a

9-30 K ey De nitions

· · underlined, the

Page 517 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 517 · click to enlarge
Page 517

directional coupler . Enter the power loss caused by the coupled

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOSS/SENSR LISTS softkey submenus described

below .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PWR RANGE AUTO man toggles the power range mode between auto and manual. A uto

mode selects the power range based on the power selected.

Manual mode limits power entry to within the selected range .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PWRMTR CAL [ ]

leads to the power meter calibration menu which provides two

types of power meter calibration, continuous and single-sample .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

turns o power meter calibration.

PWRMTR CAL [OFF]

NNNNN

R

measures the absolute power amplitude at input R.

NNNNNNNNNNNNNNNNNNNNNNNNNN

converts the active marker values into rectangular form. The

complex impedance values of the active marker are displayed in

terms of resistance , reactance , and equivalent capacitance or

inductance . This is the default Smith chart marker .

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RANGE 0 -15 TO +10 selects power range 0 when in manual power range .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

selects power range 1 when in manual power range .

RANGE 1 -25 TO 0

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RANGE 2 -35 TO -10

selects power range 2 when in manual power range .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

selects power range 3 when in manual power range .

RANGE 3 -45 TO -20

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RANGE 4 -55 TO -30

selects power range 4 when in manual power range .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

selects power range 5 when in manual power range .

RANGE 5 -65 TO -40

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

selects power range 6 when in manual power range .

RANGE 6 -75 TO -50

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RANGE 7 -85 TO -60

selects power range 7 when in manual power range .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

speci es whether or not to store the raw data (ratioed and

RAW ARRAY on OFF

averaged) on disk with the instrument state .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

selects whether sampler and attenuator o sets are ON or OFF .

RAW OFFSET On Off

By selecting raw o sets OFF , a full two port error correction

can be performed without including the e ects of the o sets .

It also saves substantial time at recalls and during frequency

changes . Raw o sets follow the channel coupling. This softkey

is used with \T ake4" mode . See \Example 2E" in Chapter 2 of

the HP 8753D Programmer's Guide.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Re/Im MKR when in the smith marker menu,

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Re/Im MKR displays the

values of the active marker on a Smith chart as a real and

imaginary pair . The complex data is separated into its real part

and imaginary part. The rst marker value given is the real

part M cos , and the second value is the imaginary part M sin

, where M = magnitude .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

When in the polar marker menu,

Re/Im MKR displays the

values of the active marker as a real and imaginary pair . The

complex data is separated into its real part and imaginary

part. The rst marker value given is the real part M cos ,

and the second value is the imaginary part M sin , where

M = magnitude .

K ey De nitions 9-31

arm with the · R+jX MKR

Page 518 of the Agilent Technologies 8753D user manual — PRESET
Page 518 · click to enlarge
Page 518

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

READ FILE TITLES

searches the directory of the disk for le names recognized

as belonging to an instrument state , and displays them in the

softkey labels . No more than ve titles are displayed at one

time . If there are more than ve , repeatedly pressing this key

causes the next ve to be displayed. If there are fewer than

ve , the remaining softkey labels are blanked.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

READ SEQ FILE TITLS

is a disk le directory command. Pressing this softkey will read

the rst six sequence titles and display them in the softkey

labels as described in Loading a Sequence When the Title Is

Not Known. These sequences can then be loaded into internal

memory .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

READ SEQ FILE TITLS is pressed again, the next six

sequence titles on the disk will be displayed. T o read the

contents of the disk starting again with the rst sequence:

remove the disk, reinsert it into the drive , and press

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

READ SEQ FILE TITLS .

NNNNNNNNNNNNNN

displays only the real (resistive) portion of the measured data

REAL

on a Cartesian format. This is similar to the linear magnitude

format, but can show both positive and negative values . It is

primarily used for analyzing responses in the time domain, and

also to display an auxiliary input voltage signal for service

purposes .

N

RECALL CAL PORT 1

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Press this key after selecting the le associated with port 1

error correction for adapter removal calibration.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECALL CAL PORT 2

Press this key after selecting the le associated with port 2

error correction for adapter removal calibration.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECALL COLORS

recalls the previously saved modi ed version of the color set.

This key appears only when a color set has been saved.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECALL KEYS MENU

provides access to the recall keys menu where speci c registers

can be recalled.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECALL KEYS on OFF presents the recall keys menu as the initial menu when

SA VE/RECALL

has been pressed.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECALL REG1

recalls the instrument state saved in register 1.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

recalls the instrument state saved in register 2.

RECALL REG2

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

recalls the instrument state saved in register 3.

RECALL REG3

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECALL REG4

recalls the instrument state saved in register 4.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

recalls the instrument state saved in register 5.

RECALL REG5

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECALL REG6

recalls the instrument state saved in register 6.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

recalls the instrument state saved in register 7.

RECALL REG7

N

RECALL STATE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

is used in conjunction with sequencing, to return the

instrument to the known preset state without turning o

the sequencing function. This is not the same as pressing

PRESET

keyno preset tests are run, and the HP-IB and

sequencing activities are not changed.

9-32 K ey De nitions

If · · the

Page 519 of the Agilent Technologies 8753D user manual — N
Page 519 · click to enlarge
Page 519

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECEIVER CAL

provides access to the Receiver Cal Menu.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

sets the position of the reference line on the graticule of a

REFERENCE POSITION

Cartesian display , with 0 the bottom line of the graticule and 10

the top line . It has no e ect on a polar or Smith display . The

reference position is indicated with a small triangle just outside

the graticule , on the left side for channel 1 and the right side

for channel 2.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REFERENCE VALUE changes the value of the reference line , moving the

measurement trace correspondingly . In polar and Smith chart

formats , the reference value is the same as the scale , and is the

value of the outer circle .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REFLFWD S11 (A/R) de nes the measurement as S

, the complex re ection

coecient (magnitude and phase) of the test device input.

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

de nes the measurement as S

, the complex re ection

REFL: REV S22 (B/R)

coecient (magnitude and phase) of the output of the device

under test.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REFLECT AND LINE measures the re ection and thru paths of the current

calibration standard.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REFLECTION leads to the re ection calibration menu.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

completes the adapter removal procedure , removing the e ects

REMOVE ADAPTER

of the adapter being used.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RENAME FILE

allows you to change the name of a le that has already been

saved.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RESET COLOR resets the color being modi ed to the default color .

NNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNN

RESPONSE When in the specify class more menu,

RESPONSE is used to

enter the standard numbers for a response calibration. This

calibration corrects for frequency response in either re ection

or transmission measurements , depending on the parameter

being measured when a calibration is performed. (F or default

kits , the standard is either the open or short for re ection

measurements , or the thru for transmission measurements .)

NNNNNNNNNNNNNNNNNNNNNNNNNN

When in the response cal menu,

RESPONSE leads to the

frequency response calibration. This is the simplest and

fastest accuracy enhancement procedure , but should be

used when extreme accuracy is not required. It e ectively

removes the frequency response errors of the test setup for

re ection or transmission measurements .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RESPONSE & ISOL'N When in the specify class more menu,

RESPONSE & ISOL'N

is used to enter the standard numbers for a response and

isolation calibration. This calibration corrects for frequency

response and directivity in re ection measurements ,

or frequency response and isolation in transmission

measurements .

When in the response and isolation menu,

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RESPONSE & ISOL'N leads to the menus used to perform

a response and isolation measurement calibration, for

measurement of devices with wide dynamic range . This

K ey De nitions 9-33

Page 520 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 520 · click to enlarge
Page 520

procedure e ectively removes the same frequency response

errors as the response calibration. In addition, it e ectively

removes the isolation (crosstalk) error in transmission

measurements or the directivity error in re ection

measurements . As well as the devices required for a simple

response calibration, an isolation standard is required.

The standard normally used to correct for isolation is an

impedance-matched load (usually 50 or 75 ohms). Response

and directivity calibration procedures for re ection and

transmission measurements are provided in the following

pages .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RESTORE DISPLAY

turns o the tabular listing and returns the measurement

display to the screen.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RESUME CAL SEQUENCE

eliminates the need to restart a calibration sequence that was

interrupted to access some other menu. This softkey goes back

to the point where the calibration sequence was interrupted.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REV ISOL'N ISOL'N STD

measures the reverse isolation of the calibration standard.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REV MATCH (Label Class)

lets you enter a label for the reverse match class . The label

appears during a calibration that uses this class .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REV MATCH (Specify Class)

speci es which standards are in the reverse match class in the

calibration kit.

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REV MATCH THRU is used to enter the standard numbers for the reverse match

(thru) calibration. (F or default kits , this is the thru.)

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REV TRANS (Label Class)

lets you enter a label for the reverse transmission class . The

label appears during a calibration that uses this class .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REV TRANS (Specify Class)

speci es which standards are in the reverse transmission class

in the calibration kit.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REV TRANS THRU is used to enter the standard numbers for the reverse

transmission (thru) calibration. (F or default kits , this is the

thru.)

N

NNNNNNNNNNNNNNNNNNNNNN

RF > LO

adjusts the source frequency higher than the LO by the amount

of the LO (within the limits of the analyzer).

NNNNNNNNNNNNNNNNNNNNNNN

RF < LO

adjusts the source frequency lower than the LO by the amount

of the LO (within the limits of the analyzer).

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RIGHT LOWER

draws a quarter-page plot in the lower right quadrant of the

page .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RIGHT UPPER

draws a quarter-page plot in the upper right quadrant of the

page .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ROUND SECONDS

resets the seconds counter to zero in real-time clock.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

S PARAMETERS

presents the S-parameter menu, which is used to de ne

the input ports and test set direction for S-parameter

measurements .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

S11 1-PORT provides a measurement calibration for re ection-only

measurements of one-port devices or properly terminated

two-port devices , at port 1 of an S-parameter test set or the test

port of a transmission/re ection test set.

9-34 K ey De nitions

Page 521 of the Agilent Technologies 8753D user manual — N
Page 521 · click to enlarge
Page 521

NNNNNNNNNNNNNN

S11A

is used to enter the standard numbers for the rst class

required for an S

1-port calibration. (F or default cal kits , this

is the open.)

NNNNNNNNNNNNNN

S11B

is used to enter the standard numbers for the second class

required for an S

1-port calibration. (F or default cal kits , this

is the short.)

NNNNNNNNNNNNNN

S11C

is used to enter the standard numbers for the third class

required for an S

1-port calibration. (F or default kits , this is

the load.)

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

S11 REFL SHORT

measures the short circuit TRL/LRM calibration data for

PORT 1.

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

S22 1-PORT provides a measurement calibration for re ection-only

measurements of one-port devices or properly terminated

two-port devices , at port 2 of an S-parameter test set or the test

port of a transmission/re ection test set.

NNNNNNNNNNNNNN

S22A

is used to enter the standard numbers for the rst class

required for an S

1-port calibration. (F or default cal kits , this

is the open.)

NNNNNNNNNNNNNN

S22B

is used to enter the standard numbers for the second class

required for an S

1-port calibration. (F or default cal kits , this

is the short.)

NNNNNNNNNNNNNN

S22C

is used to enter the standard numbers for the third class

required for an S

1-port calibration. (F or default kits , this is

the load.)

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

S22 REFL SHORT

measures the short circuit TRL/LRM calibration data for

PORT 2.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SAMPLR COR on OFF selects whether sampler correction is on or o .

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

saves the modi ed version of the color set.

SAVE COLORS

provides access to all the menus used for saving and recalling

SA VE/RECALL

instrument states in internal memory and for storing to , or

loading from, external disk. This includes the menus used to

de ne titles for internal registers and external disk les , to

de ne the content of disk les , to initialize disks for storage ,

and to clear data from the registers or purge les from disk.

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

stores the user-modi ed or user-de ned kit into memory , after

SAVE USER KIT

it has been modi ed.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SAVE USING ASCII selects ASCII format for data storage to disk.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

selects binary format for data storage .

SAVE USING BINARY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SCALE/DIV

changes the response value scale per division of the displayed

trace . In polar and Smith chart formats , this refers to the full

scale value at the outer circumference , and is identical to

reference value .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SCALE PLOT [ ] toggles between two selections for plot scale , FULL and GRA T .

N

SCALE PLOT [FULL]

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

is the normal scale selection for plotting on blank paper . It

includes space for all display annotations such as marker

values , stimulus values , etc . The entire display ts within the

K ey De nitions 9-35

Page 522 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 522 · click to enlarge
Page 522

user-de ned boundaries of P1 and P2 on the plotter , while

maintaining the exact same aspect ratio as the display .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SCALE PLOT [GRAT]

expands or reduces the horizontal and vertical scale so that the

lower left and upper right graticule corners exactly correspond

to the user-de ned P1 and P2 scaling points on the plotter . This

is convenient for plotting on preprinted rectangular or polar

forms (for example , on a Smith Chart).

SCALE REF

] makes scale per division the active function. A menu is

displayed that is used to modify the vertical axis scale and

the reference line value and position. In addition this menu

provides electrical delay o set capabilities for adding or

subtracting linear phase to maintain phase linearity .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEARCH LEFT

searches the trace for the next occurrence of the target value

to the left.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEARCH RIGHT searches the trace for the next occurrence of the target value

to the right.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEARCH: MAX

moves the active marker to the maximum point on the trace .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

moves the active marker to the minimum point on the trace .

SEARCH: MIN

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEARCH: OFF

turns o the marker search function.

NNNNNNNNNNNNNNNNNNNNNNN

SEGMENT

speci es which limit segment in the table is to be modi ed.

A maximum of three sets of segment values are displayed at

one time , and the list can be scrolled up or down to show

other segment entries . Use the entry block controls to move

the pointer > to the required segment number . The indicated

segment can then be edited or deleted. If the table of limits is

designated \EMPTY ," new segments can be added using the

NNNNNNNNNNN

NNNNNNNNNNNNNN

EDIT softkey .

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEGMENT: CENTER

sets the center frequency of a subsweep in a list frequency

sweep .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEGMENT: SPAN

sets the frequency or power span of a subsweep about a

speci ed center frequency .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEGMENT: START

sets the start frequency of a subsweep .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEGMENT: STOP

sets the stop frequency of a subsweep .

NNNNNNNNNNNNNNNNNNNNNNNNNN

leads to the select quadrant menu, which provides the

SEL QUAD

capability of drawing quarter-page plots . This is not used for

printing.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

provides access to the select disk menu.

SELECT DISK

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SELECT LETTER

The active entry area displays the letters of the alphabet, digits

0 through 9, and mathematical symbols . T o de ne a title , rotate

the knob until the arrow " points at the rst letter , then press

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SELECT LETTER . Repeat this until the complete title is de ned,

for a maximum of 50 characters . As each character is selected,

it is appended to the title at the top of the graticule .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SELF DIAGNOSE

prompts the analyzer to run a series of tests to determine a

problem.

9-36 K ey De nitions

· ADD or

Page 523 of the Agilent Technologies 8753D user manual — SET Z0
Page 523 · click to enlarge
Page 523

SEQ

accesses a series of sequencing menus . These allow you to

create , modify , and store up to 6 sequences which can be run

automatically .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQUENCE 1 SEQ1

activates editing mode for the segment titled \SEQ1" (default

title).

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQUENCE 2 SEQ2

activates editing mode for the segment titled \SEQ2" (default

title).

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQUENCE 3 SEQ3

activates editing mode for the segment titled \SEQ3" (default

title).

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQUENCE 4 SEQ4

activates editing mode for the segment titled \SEQ4" (default

title).

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQUENCE 5 SEQ5

activates editing mode for the segment titled \SEQ5" (default

title).

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQUENCE 6 SEQ6

activates editing mode for the segment titled \SEQ6" (default

title).

N

SEQUENCE FILENAMING

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

accesses a lenaming menu which is used to automatically

increment or decrement the name of a le that is generated by

the network analyzer during a SEQUENCE.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SERVICE MENU

leads to a series of service and test menus described in detail in

the On-Site System Service Manual.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SERVICE MODES

a collection of common modes used for troubleshooting.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

goes to the address menu, which is used to set the HP-IB

SET ADDRESSES

address of the analyzer , and to display and modify the

addresses of peripheral devices in the system.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

allows you to set the analyzer's internal clock.

SET CLOCK

NNNNNNNNNNNNNNNNNNNNNNN

SET DAY

allows you to set the day in the analyzer's internal clock.

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

(Option 010 only) changes the frequency sweep to harmonic

SET FREQ LOW PASS

intervals to accommodate time domain low-pass operation

(option 010). If this mode is used, the frequencies must be set

before calibration.

NNNNNNNNNNNNNNNNNNNNNNNNNN

SET HOUR allows you to set the hour in the analyzer's internal clock.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

allows you to set the minutes in the analyzer's internal clock.

SET MINUTES

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SET MONTH

allows you to set the month in the analyzer's internal clock.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

sets the measurement reference plane to the TRL/LRM

SET REF: REFLECT

REFLECT standard.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SET REFTHRU sets the measurement reference plane to the TRL/LRM THRU

standard.

NNNNNNNNNNNNNNNNNNNNNNNNNN

SET YEAR allows you to set the year in the analyzer's internal clock.

NNNNNNNNNNNNNNNNNNNN

sets the characteristic impedance used by the analyzer in

SET Z0

calculating measured impedance with Smith chart markers and

conversion parameters . Characteristic impedance must be set

correctly before calibration procedures are performed.

K ey De nitions 9-37

Page 524 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 524 · click to enlarge
Page 524

NNNNNNNNNNNNNNNNN

SHORT

de nes the standard type as a short, for calibrating re ection

measurements . Shorts are assigned a terminal impedance of

0 ohms , but delay and loss o sets may still be added.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SHOW MENUS

used to display a speci c menu prior to a pause statement.

NNNNNNNNNNNNNNNNNNNN

takes one sweep of data and returns to the hold mode .

SINGLE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

sets the limits at a single stimulus point. If limit lines are on,

SINGLE POINT

the upper limit value of a single point limit is displayed as n 9,

and the lower limit is displayed as n 8. A limit test at a single

point not terminating a at or sloped line tests the nearest

actual measured data point. A single point limit can be used as

a termination for a at line or sloping line limit segment. When

a single point terminates a sloping line or when it terminates a

at line and has the same limit values as the at line , the single

point is not displayed as n 9 and n 8. The indication for a sloping

line segment in the displayed table of limits is SP .

N

SINGLE SEG SWEEP

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

enables a measurement of a single segment of the frequency

list, without loss of calibration. The segment to be measured is

selected using the entry block.

In single segment mode , selecting a measurement calibration

will force the full list sweep before prompting for calibration

standards . The calibration will then be valid for any single

segment.

If an instrument state is saved in memory with a single-segment

trace , a recall will re-display that segment while also recalling

the entire list.

N

NNNNNNNNNNNNNNNNNNNNNN

SLIDING de nes the load as a sliding load. When such a load is measured

during calibration, the analyzer will prompt for several load

positions , and calculate the ideal load value from it.

NNNNNNNNNNNNNNNNN

SLOPE

compensates for power loss versus the frequency sweep , by

sloping the output power upwards proportionally to frequency .

Use this softkey to enter the power slope in dB per GHz of

sweep .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

toggles the power slope function on or o . With slope on, the

output power increases with frequency , starting at the selected

power level.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SLOPING LINE

de nes a sloping limit line segment that is linear with frequency

or other stimulus value , and is continuous to the next stimulus

value and limit. If a sloping line is the nal segment it becomes

a at line terminated at the stop stimulus . A sloping line

segment is indicated as SL on the displayed table of limits .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SMITH CHART

displays a Smith chart format. This is used in re ection

measurements to provide a readout of the data in terms of

impedance .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SMITH MKR MENU

leads to a menu of special markers for use with a Smith chart

format.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SMOOTHING APERTURE lets you change the value of the smoothing aperture as a

percent of the span. When smoothing aperture is the active

9-38 K ey De nitions

SLOPE on OFF

Page 525 of the Agilent Technologies 8753D user manual — ST ART
Page 525 · click to enlarge
Page 525

function, its value in stimulus units is displayed below its

percent value in the active entry area.

Smoothing aperture is also used to set the aperture for group

delay measurements . Note that the displayed smoothing

aperture is not the group delay aperture unless smoothing is on.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

turns the smoothing function on or o for the active channel.

When smoothing is on, the annotation \Smo" is displayed in the

status notations area.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SOURCE PWR on OFF turns the source power on or o . Use this key to restore

NNNNNNNNNNNNNNNNN

power after a power trip has occurred. (See the

description.)

NNNNNNNNNNNNNNNNN

SPACE

inserts a space in the title .

is used, along with the

key , to de ne the frequency

SP AN

CENTER

range of the stimulus . When the

key is pressed it

SP AN

becomes the active function. The value is displayed in the

active entry area, and can be changed with the knob , step keys ,

or numeric keypad.

NNNNNNNNNNNNNN

sets the frequency or power span of a subsweep about a

SPAN

speci ed center frequency .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

presents the special function menu.

SPECIAL FUNCTIONS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SPECIFY CLASS

leads to the specify class menu. After the standards are

modi ed, use this key to specify a class to consist of certain

standards .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SPECIFY CLASS DONE nishes the specify class function and returns to the modify cal

kit menu.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SPECIFY GATE (Option 010 only) is used to specify the parameters of the gate .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SPECIFY OFFSET

allows additional speci cations for a user-de ned standard.

F eatures speci ed in this menu are common to all ve types of

standards .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SPLIT DISP on OFF toggles between a full-screen single graticule display of one

or both channels , and a split display with two half-screen

graticules one above the other . The split display can be

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

used in conjunction with

DUAL CH ON in the display menu

to show the measured data of each channel simultaneously

on separate graticules . In addition, the stimulus functions

of the two channels can be controlled independently using

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

COUPLED CH ON in the stimulus menu. The markers can

also be controlled independently for each channel using

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKERS: UNCOUPLED in the marker mode menu.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

selects whether spur avoidance is ON or OFF . Selecting spur

SPUR AVOID On Off

avoidance OFF , along with selecting raw o sets OFF , saves

substantial time at recalls and during frequency changes . Spur

avoidance is always coupled between channels .

ST ART

is used to de ne the start frequency of a frequency range .

ST ART

key is pressed it becomes the active function.

K ey De nitions 9-39

Show 5 more labels / short text on this page
SMOOTHING on OFF · POWER key · · · When the
Page 526 of the Agilent Technologies 8753D user manual — STIMULUS VALUE
Page 526 · click to enlarge
Page 526

The value is displayed in the active entry area, and can be

changed with the knob , step keys , or numeric keypad.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

calculates and displays the mean, standard deviation, and

peak-to-peak values of the section of the displayed trace

between the active marker and the delta reference marker .

If there is no delta reference , the statistics are calculated for

the entire trace . A convenient use of this feature is to nd

the peak-to-peak value of passband ripple without searching

separately for the maximum and minimum values .

The statistics are absolute valuesthe delta marker here serves

to de ne the span. F or polar and Smith chart formats the

statistics are calculated using the rst value of the complex pair

(magnitude , real part, resistance , or conductance).

After each standard is de ned, including o sets , press

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STD DONE (DEFINED) to terminate the standard de nition.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

returns to the de ne standard menu.

STD DONE (DEFINED)

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STD OFFSET DONE

is used to end the specify o set sequence .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

is used to specify the type of calibration device being measured.

STD TYPE:

STD TYPE:

de nes the standard type to be a load, but with an arbitrary

ARBITRARY IMPEDANCE

impedance (di erent from system Z0).

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

de nes the standard type as a transmission line of speci ed

STD TYPE: DELAY/THRU

length, for calibrating transmission measurements .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

de nes the standard type as a load (termination). Loads

STD TYPE: LOAD

are assigned a terminal impedance equal to the system

characteristic impedance ZO , but delay and loss o sets may still

be added. If the load impedance is not ZO , use the arbitrary

impedance standard de nition.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

de nes the standard type as an open used for calibrating

STD TYPE: OPEN

re ection measurements . Opens are assigned a terminal

impedance of in nite ohms , but delay and loss o sets may still

be added. Pressing this key also brings up a menu for de ning

the open, including its capacitance .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

de nes the standard type as a short used for calibrating

STD TYPE: SHORT

re ection measurements . Shorts are assigned a terminal

impedance of 0 ohms , but delay and loss o sets may still be

added.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STEP SIZE is used to specify the subsweep in frequency steps instead

of number of points . Changing the start frequency , stop

frequency , span, or number of points may change the step size .

Changing the step size may change the number of points and

stop frequency in start/stop/step mode; or the frequency span

in center/span/step mode . In each case , the frequency span

becomes a multiple of the step size .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

sets the starting stimulus value of a segment, using entry block

STIMULUS VALUE

controls . The ending stimulus value of the segment is de ned

by the start of the next line segment. No more than one

segment can be de ned over the same stimulus range .

9-40 K ey De nitions

Show 60 more labels / short text on this page
STATS on OFF · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ·
Page 527 of the Agilent Technologies 8753D user manual — TALKER/LISTENER
Page 527 · click to enlarge
Page 527

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STIMULUS OFFSET

adds or subtracts an o set in stimulus value . This allows limits

already de ned to be used for testing in a di erent stimulus

range . Use the entry block controls to specify the o set

required.

STOP

is used to de ne the stop frequency of a frequency range .

STOP

key is pressed it becomes the active function.

The value is displayed in the active entry area, and can be

changed with the knob , step keys , or numeric keypad.

NNNNNNNNNNNNNN

STOP

sets the stop frequency of a subsweep .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

presents the store sequence to disk menu with a list of

STORE SEQ TO DISK

sequences that can be stored.

NNNNNNNNNNNNNNNNN

is used to set the frequency of the LO source to sweep .

SWEEP

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SWEEP TIME [ ]

toggles between automatic and manual sweep time .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

presents the sweep type menu, where one of the available

SWEEP TYPE MENU

types of stimulus sweep can be selected.

N

NNNNNNNNNN

reformats a re ection measurement into its equivalent SWR

SWR

(standing wave ratio) value . SWR is equivalent to (1+)/(1),

where is the re ection coecient. Note that the results are

valid only for re ection measurements . If the SWR format is

used for measurements of S

the results are not valid.

presents the system menu.

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

is the mode used when peripheral devices are to be used and

SYSTEM CONTROLLER

there is no external controller . In this mode , the analyzer can

directly control peripherals (plotter , printer , disk drive , or

power meter). System controller mode must be set in order for

the analyzer to access peripherals from the front panel to plot,

print, store on disk, or perform power meter functions , if there

is no other controller on the bus .

The system controller mode can be used without knowledge

of HP-IB programming. However , the HP-IB address must be

entered for each peripheral device .

This mode can only be selected manually from the analyzer's

front panel, and can be used only if no active computer

controller is connected to the system through HP-IB . If you

try to set system controller mode when another controller

is present, the message ANOTHER SYSTEM CONTROLLER ON

HP-IB BUS is displayed. Do not attempt to use this mode for

programming.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TAKE CAL SWEEP

Each data point is measured during the initial sweep and the

correction data is placed in the power meter correction table .

This provides data usable in the

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ONE SWEEP mode .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TAKE RCVR CAL SWEEP

executes a receiver calibration.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

is the mode normally used for remote programming of the

TALKER/LISTENER

analyzer . In this mode , the analyzer and all peripheral devices

are controlled from the external controller . The controller can

command the analyzer to talk, and the plotter or other device

to listen. The analyzer and peripheral devices cannot talk

K ey De nitions 9-41

Show 4 more labels / short text on this page
· When the · or S ·
Page 528 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNN
Page 528 · click to enlarge
Page 528

directly to each other unless the computer sets up a data path

between them.

This mode allows the analyzer to be either a talker or a listener ,

as required by the controlling computer for the particular

operation in progress .

A talker is a device capable of sending out data when it is

addressed to talk. There can be only one talker at any given

time . The analyzer is a talker when it sends information over

the bus .

A listener is a device capable of receiving data when it is

addressed to listen. There can be any number of listeners at

any given time . The analyzer is a listener when it is controlled

over the bus by a computer .

N

NNNNNNNNNNNNNNNNNNN

TARGET

makes target value the active function, and places the active

marker at a speci ed target point on the trace . The default

target value is 3 dB . The target menu is presented, providing

search right and search left options to resolve multiple

solutions .

F or relative measurements , a search reference must be de ned

with a delta marker or a xed marker before the search is

activated.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TERMINAL IMPEDANCE is used to specify the (arbitrary) impedance of the standard, in

ohms .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TEST OPTIONS is used to set con gurations before running the service tests .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TESTPORT 1 2

is used to direct the RF power to port 1 or port 2. (F or non-S

parameter inputs only .)

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TESTSET I/O FWD

is used to support specialized test sets , such as a testset that

measures duplexers . It allows you to set three bits (D1, D2,

and D3) to a value of 0 to 7, and outputs it as binary from the

rear panel testset connector . It tracks the coupling ag, so if

coupling is on, and FWD channel 1 is the active channel, FWD

channel 2 will be set to the same value .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TESTSET I/O REV is used to support specialized testsets , such as a testset that

measures duplexers . It allows you to set three bits (D1, D2,

and D3) to a value of 0 to 7, and outputs it as binary from the

rear panel testset connector . It tracks the coupling ag, so if

coupling is on, and REV channel 1 is the active channel, REV

channel 2 will be set to the same value .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TESTSET SW XXXX toggles the internal solid state switch from a hold mode , to a

continuously switching mode , or to a number of sweeps mode

when full 2-port correction is enabled. Use for fast 2-port

calibration.

N

NNNNNNNNNNNNNNNN

TESTS presents the service test menu.

NNNNNNNNNNNNNN

TEXT

selects all the non-data text for color modi cation. F or

example: operating parameters .

NNNNNNNNNNNNNNNNNNNNNNNNNN

TEXT [ ]

brings up the color de nition menu. The default color for text

is black.

9-42 K ey De nitions

Page 529 of the Agilent Technologies 8753D user manual — LOCAL
Page 529 · click to enlarge
Page 529

NNNNNNNNNNNNNN

THRU a calibration standard type .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

measures all four S-parameters in a TRL/LRM calibration.

THRU THRU

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

turns the time stamp function on or o .

TIME STAMP on OFF

NNNNNNNNNNNNNN

TINT

adjusts the continuum of hues on the color wheel of the chosen

attribute . See A djusting Color for an explanation of using this

softkey for color modi cation of display attributes .

NNNNNNNNNNNNNNNNN

TITLE

presents the title menu in the softkey labels area and the

character set in the active entry area. These are used to label

the active channel display . A title more menu allows up to four

values to be included in the printed title; active entry , active

marker amplitude , limit test results , and loop counter value .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TITLE SEQUENCE

allows the operator to rename any sequence with an eight

character title . All titles entered from the front panel must

begin with a letter , and may only contain letters and numbers .

A procedure for changing the title of a sequence is provided at

the beginning of this chapter .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TITLE TO MEMORY

moves the title string data obtained with the

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

P MTR/HPIB TO TITLE command into a data array .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TITLE TO MEMORY strips o leading characters that are not

numeric , reads the numeric value , and then discards everything

else . The number is converted into analyzer internal format,

and is placed into the real portion of the memory trace at:

Display point = total points - 1 - loop counter

If the value of the loop counter is zero , then the title number

goes in the last point of memory . If the loop counter is greater

than or equal to the current number of measurement points ,

the number is placed in the rst point of memory . A data to

memory command must be executed before using the title to

memory command.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TITLE TO P MTR/HPIB

outputs a title string to any device with an HP-IB address

that matches the address set with the analyzer

LOCAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SET ADDRESSES

ADDRESSP MTR/HPIB commands . This

softkey is generally used for two purposes:

Sending a title to a printer when a CR-LF is not desired.

Sending commands to an HP-IB device .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TITLE TO PERIPHERAL

outputs a title string to any device with an HP-IB address

that matches the address set with the analyzer

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SPECIAL FUNCTIONS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PERIPHERAL HPIB ADDR commands . This

softkey is generally used for two purposes:

Sending a title to a printer when a CR-LF is not desired.

Sending commands to an HP-IB device .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

outputs a title string to any device with an HP-IB address

TITLE TO PRNTR/HPIB

that matches the address set with the analyzer

LOCAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SET ADDRESSES

ADDRESSPRINTER commands . This softkey

is generally used for two purposes:

K ey De nitions 9-43

· ·

Page 530 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 530 · click to enlarge
Page 530

Sending a title to a printer for data logging or documentation

purposes .

Sending commands to a printer or other HP-IB device .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

is used in conjunction with other search features to track the

search with each new sweep . Turning tracking on makes the

analyzer search every new trace for the speci ed target value

and put the active marker on that point. If bandwidth search

is on, tracking searches every new trace for the speci ed

bandwidth, and repositions the dedicated bandwidth markers .

When tracking is o , the target is found on the current sweep

and remains at the same stimulus value regardless of changes in

trace response value with subsequent sweeps .

A maximum and a minimum point can be tracked

simultaneously using two channels and uncoupled markers .

N

TRANS DONE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

goes back to the two-port cal menu when transmission

measurements are nished.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRANSFWD S21 (B/R) de nes the measurement as S

, the complex forward

transmission coecient (magnitude and phase) of the test

device .

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRANSREV S12 (A/R) de nes the measurement as S

, the complex reverse

transmission coecient (magnitude and phase) of the test

device .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRANSFORM MENU (Option 010 only) leads to a series of menus that transform the

measured data from the frequency domain to the time domain.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

(Option 010 only) switches between time domain transform on

and o .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRANSMISSION leads to the transmission menu.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRIGGER MENU

presents the trigger menu, which is used to select the type and

number of the sweep trigger .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRIGGER: TRIG OFF

turns o external trigger mode .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

leads to the TRL*/LRM* 2-port calibration menu.

TRL*/LRM* 2-PORT

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRL/LRM OPTION

selects the TRL/LRM Option Menu.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

is used to enter the standard numbers for the TRL LINE or

TRL LINE OR MATCH

MA TCH class .

NNNNNNNNNNNNNNNNNNNNNNNNNN

TRL THRU is used to enter the standard numbers for the TRL THRU class .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRL REFLECT

is used to enter the standard numbers for the TRL REFLECT

class .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TTL OUT HIGH sets the TTL output (TEST SEQ BNC) on the back of the

analyzer high.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TTL OUT LOW

sets the TTL output (TEST SEQ BNC) on the back of the

analyzer low .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TUNED RECEIVER

sets the analyzer to function as a tuned receiver only , disabling

the source .

9-44 K ey De nitions

TRACKING on OFF · TRANSFORM on OFF

Page 531 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 531 · click to enlarge
Page 531

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

UNCOUPLED allows the marker stimulus values to be controlled

independently on each channel.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

sends the sum frequency of the RF and LO to the R channel.

UP CONVERTER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

UPPER LIMIT

sets the upper limit value for the start of the segment. If a

lower limit is speci ed, an upper limit must also be de ned. If

no upper limit is required for a particular measurement, force

the upper limit value out of range (for example +500 dB).

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOWER LIMIT is pressed, all the

segments in the table are displayed in terms of upper and lower

limits , even if they were de ned as delta limits and middle

value .

If you attempt to set an upper limit that is lower than the lower

limit, or vice versa, both limits will be automatically set to the

same value .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

USE MEMORY ON off (Option 010 only) remembers a speci ed window pulse width

(or step rise time) di erent from the standard window values .

A window is activated only for viewing a time domain response ,

and does not a ect a displayed frequency domain response .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

USE PASS CONTROL

lets you control the analyzer with the computer over HP-IB as

with the talker/listener mode , and also allows the analyzer to

become a controller in order to plot, print, or directly access

an external disk. During this peripheral operation, the host

computer is free to perform other internal tasks that do not

require use of the bus (the bus is tied up by the network

analyzer during this time).

The pass control mode requires that the external controller is

programmed to respond to a request for control and to issue

a take control command. When the peripheral operation is

complete , the analyzer passes control back to the computer .

Refer to the \HP-IB Programming Reference" and \HP-IB

Programming Examples" chapters in the HP 8753D Network

Analyzer Programmer's Guide for more information.

In general, use the talker/listener mode for programming the

analyzer unless direct peripheral access is required.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

USE SENSOR A/B

selects the A or B power sensor calibration factor list for use in

power meter calibration measurements .

NNNNNNNNNNNNNN

USER

is used to select the preset condition de ned by the user .

NNNNNNNNNNNNNNNNNNNNNNNNNN

is used to de ne kits other than those o ered by

USER KIT

Hewlett-P ackard.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

VELOCITY FACTOR Enters the velocity factor used by the analyzer to calculate

equivalent electrical length in distance-to-fault measurements

using the time domain option. V alues entered should be less

than 1.

V elocity factor is the ratio of the velocity of wave propagation

in a coaxial cable to the velocity of wave propagation in free

space . Most cables have a relative velocity of about 0.66 the

K ey De nitions 9-45

When · UPPER LIMIT or

Page 532 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 532 · click to enlarge
Page 532

speed in free space . This velocity depends on the relative

permittivity of the cable dielectric (

V el ocity F actor =

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

VIEW MEASURE

toggles to become view setup when the analyzer is in frequency

o set mode .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

VOLUME NUMBER speci es the number of the disk volume to be accessed. In

general, all 3.5 inch oppy disks are considered one volume

(volume 0). F or hard disk drives , such as the HP 9153A

(Winchester), a switch in the disk drive must be set to de ne

the number of volumes on the disk. F or more information,

refer to the manual for the individual hard disk drive .

NNNNNNNNNNNNNNNNNNNN

WAIT x pauses the execution of subsequent sequence commands for x

number of seconds . T erminate this command with

.

Entering a 0 in wait x causes the instrument to wait for prior

sequence command activities to nish before allowing the

next command to begin. The wait 0 command only a ects

the command immediately following it, and does not a ect

commands later in the sequence .

NNNNNNNNNNNNNNNNNNNNNNN

WARNING

selects the warning annotation for color modi cation.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

brings up the color de nition menu. The warning annotation

WARNING [ ]

default color is black.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

WAVEGUIDE de nes the standard (and the o set) as rectangular waveguide .

This causes the analyzer to assume a dispersive delay (see

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

OFFSET DELAY above).

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

WAVEGUIDE DELAY

applies a non-linear phase shift for use with electrical delay

which follows the standard dispersive phase equation for

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

rectangular waveguide . When

WAVEGUIDE DELAY is pressed,

the active function becomes the W A VEGUIDE CUTOFF

frequency , which is used in the phase equation. Choosing a

Start frequency less than the Cuto frequency results in phase

errors .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

WIDTH VALUE is used to set the amplitude parameter (for example 3 dB) that

de nes the start and stop points for a bandwidth search. The

bandwidth search feature analyzes a bandpass or band reject

trace and calculates the center point, bandwidth, and Q (quality

factor) for the speci ed bandwidth. Bandwidth units are the

units of the current format.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

WIDTHS on OFF turns on the bandwidth search feature and calculates the

center stimulus value , bandwidth, and Q of a bandpass or band

reject shape on the trace . The amplitude value that de nes the

passband or rejectband is set using the

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

WIDTH VALUE softkey .

F our markers are turned on, and each has a dedicated use .

Marker 1 is a starting point from which the search is begun.

Marker 2 goes to the bandwidth center point. Marker 3 goes to

the bandwidth cuto point on the left, and Marker 4 to the

cuto point on the right.

9-46 K ey De nitions

Show 5 more labels / short text on this page
r · ) as · p · " · r
Page 533 of the Agilent Technologies 8753D user manual — Z: REFL
Page 533 · click to enlarge
Page 533

If a delta marker or xed marker is on, it is used as the

reference point from which the bandwidth amplitude is

measured. F or example , if marker 1 is the delta marker and is

set at the passband maximum, and the width value is set to

3 dB , the bandwidth search nds the bandwidth cuto points

3 dB below the maximum and calculates the 3 dB bandwidth

and Q.

If marker 2 (the dedicated bandwidth center point marker) is

the delta reference marker , the search nds the points 3 dB

down from the center .

If no delta reference marker is set, the bandwidth values are

absolute values .

NNNNNNNNNNNNNNNNNNNN

WINDOW (Option 010 only) is used to specify the parameters of the

window in the transform menu.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

WINDOWMAXIMUM (Option 010 only) sets the pulse width to the widest value

allowed. This minimizes the sidelobes and provides the greatest

dynamic range .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

WINDOW: MINIMUM

(Option 010 only) is used to set the window of a time domain

measurement to the minimum value . Provides essentially no

window .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

WINDOWNORMAL (Option 010 only) is used to set the window of a time domain

measurement to the normal value . Usually the most useful

because it reduces the sidelobes of the measurement somewhat.

is used to terminate basic units: dB , dBm, Hz, dB/GHz, degrees ,

or seconds . It may also be used to terminate unitless entries

such as averaging factor .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

XMIT CNTRL [ ]

toggles the PLOTTER/PRINTER serial port data transmit

control mode between the Xon-Xo protocol handshake and

the DTR-DSR (data terminal ready-data set ready) hardwire

handshake .

NNNNNNNNNNNNNNNNNNNNNNN

Y: REFL converts re ection data to its equivalent admittance values .

NNNNNNNNNNNNNNNNNNNNNNNNNN

Y: TRANS

converts transmission data to its equivalent admittance values .

NNNNNNNNNNNNNNNNNNNNNNN

converts re ection data to its equivalent impedance values .

Z: REFL

NNNNNNNNNNNNNNNNNNNNNNNNNN

Z: TRANS

converts transmission data to its equivalent impedance values .

K ey De nitions 9-47

Page 534 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 534 · click to enlarge
Page 534

Cross Reference of K ey Function to Programming Command

The following table lists the front-panel keys and softkeys alphabetically . The \Command"

column identi es the command that is similar to the front-panel or softkey function. Softkeys

that do not have corresponding programming commands are not included in this section.

T able 9-1. Cross Reference of K ey Function to Programming Command

K ey Name Command

Step Up UP

Step Down DOWN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MODE OFF Delta Marker Mode O DELO

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REF = 1 Delta Reference = Marker 1 DELR1

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REF = 2 Delta Reference = Marker 2 DELR2

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REF = 3 Delta Reference = Marker 3 DELR3

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REF = 4 Delta Reference = Marker 4 DELR4

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REF = 5 Delta Reference = Marker 5 DELR5

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REF = FIXED MKR Delta Reference = Delta Fixed

DELRFIXM

NNNNNNNNNNN

1/S Inverted S-P arameters CONVIDS

NNNNN

A Measure Channel A MEASA

NNNNNNNNNNN

A/B Ratio of A to B AB

NNNNNNNNNNN

A/R Ratio of A to R AR

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ADAPTER:COAX A dapter:Coax ADPTCO AX

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ADAPTER:WAVEGUIDE A dapter:W aveguide ADPTW A VE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ADAPTER DELAY A dapter Delay AD AP1

NNNNNNNNNNN

ADD A dd SADD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ADDRESSCONTROLLER A ddress of Controller ADDRCONT

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ADDRESSDISK A ddress of Disk ADDRDISC

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ADDRESSP MTR/HPIB A ddress of P ower Meter/HPIB ADDRPOWM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ALL SEGS SWEEP All Segments Sweep ASEG

N

ALTERNATE A and B Alternate A and B AL T AB

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AMPLITUDE OFFSET Amplitude O set LIMIAMPO

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ANALOG BUS ON off Analog Bus On ANAB

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ANALOG IN Aux Input Analog In ANAI

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ARBITRARY IMPEDANCE Arbitrary Impedance STDT ARBI

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ASSERT SRQ Service Request ASSS

9-48 K ey De nitions

Show 4 more labels / short text on this page
· · · Marker
Page 535 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 535 · click to enlarge
Page 535

T able 9-1.

Cross Reference of K ey Function to Programming Command (continued)

K ey Name Command

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AUTO FEED ON off (Plotter) Plotter A uto F eed On PL TTRA UTFON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AUTO FEED on OFF (Plotter) Plotter A uto F eed O PL TTRA UTFOFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AUTO FEED ON off (Printer) Printer A uto F eed On PRNTRA UTFON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AUTO FEED on OFF (Printer) Printer A uto F eed O PRNTRA UTOFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AUTO SCALE A uto Scale A UTO

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AVERAGING FACTOR A veraging F actor A VERF A CT

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AVERAGING ON off A veraging On A VERON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AVERAGING on OFF A veraging O A VEROFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AVERAGING RESTART A veraging Restart A VERREST

A V G

A verage MENU A VG

NNNNN

B Measure Channel B MEASB

NNNNNNNNNNN

B/R Ratio of B to R BR

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

BACKGROUND INTENSITY Background Intensity B A CI

NNNNNNNNNNNNNNNNNNNNNNNNNN

BANDPASS Bandpass B ANDP ASS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

BEEP DONE ON off Beep Done On BEEPDONEON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

BEEP DONE on OFF Beep Done O BEEPDONEOFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

BEEP FAIL ON off Beep F ail On BEEPF AILON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

BEEP FAIL on OFF Beep F ail O BEEPF AILOFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

BEEP WARN ON off Beep W arn On BEEPW ARNON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

BEEP WARN on OFF Beep W arn O BEEPW ARNOFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

BLANK DISPLAY Blank Display On BLADON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNN

C0 C0 T erm C0

NNNNNNNN

C1 C1 T erm C1

NNNNNNNN

C2 C2 T erm C2

NNNNNNNN

C3 C3 T erm C3

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL FACTOR Calibration F actor CALFCALF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL FACTOR SENSOR A Calibration F actor Sensor A CALFSENA

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL FACTOR SENSOR B Calibration F actor Sensor B CALFSENB

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL KIT2.4mm 2.4mm Calibration Kit CALK24MM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL KIT2.92* 2.92* Calibration Kit CALK292S

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL KIT2.92mm 2.92mm Calibration Kit CALK292MM

K ey De nitions 9-49

Show 4 more labels / short text on this page
· BRIGHTNESS Brightness CBRI · · Calibrate MENUCAL
Page 536 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNN
Page 536 · click to enlarge
Page 536

T able 9-1.

Cross Reference of K ey Function to Programming Command (continued)

K ey Name Command

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL KIT3.5mmC 3.5mmC Calibration Kit CALK35MC

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL KIT3.5mmD 3.5mmD Calibration Kit CALK35MD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL KITTRL 3.5mm TRL 3.5mm Calibration Kit CALKTRLK

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL KIT7mm 7mm Calibration Kit CALK7MM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL KITN 50 Type-N 50 Calibration Kit CALKN50

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL KITN 75 Type-N 75 Calibration Kit CALKN75

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL KITUSER KIT User Calibration Kit CALKUSED

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL ZOLINE ZO line impedance CALZINE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL ZOSYSTEM ZO System impedance CALZSYST

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CALIBRATENONE Calibrate None CALN

NNNNNNNNNNNNNNNNNNNN

CENTER Center , list freq subsweep CENT

CHAN 1 Channel 1 A ctive CHANT

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CH1 DATA [ ] Channel 1 Data (Color) PCOLD A T A1

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CH1 DATA LIMIT LN Channel 1 Data/Limit Line COLOCH1D

NNNNNNNNNNNNNNNNNNNNNNN

CH1 MEM Channel 1 Memory COLOCH1M

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CH1 MEM [ ] Channel 1 Memory (Color) PCOLMEMO1

CHAN 2 Channel 2 A ctive CH2

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CH2 DATA [ ] Channel 2 Data (Color) PCOLD A T A2

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CH2 DATA LIMIT LN Channel 2 Data/Limit Line COLOCH2D

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CH2 MEM [ ] Channel 2 Memory (Color) PCOLMEMO2

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CH2 MEM REF LINE Channel 2 Memory Reference Line COLOCH2M

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CH PWR [COUPLED] Channel P ower Coupled CHANPCPLD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CH PWR [UNCOUPLED] Channel P ower Uncoupled CHANPUNCPLD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CHOP A and B Chop A and B CHOP AB

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CLASS DONE Class Done CLAD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CLEAR BIT Clear Bit CLEABIT

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CLEAR LIST Clear List CLEAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CLEAR SEQUENCE Clear Sequence CLEASEn

NNNNNNNNNNNNNN

COAX Coax CO AX

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

COAXIAL DELAY Coaxial Delay CO AD

NNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CONTINUE SEQUENCE Continue Sequence CONS

1 CALK35MM selects the HP 85053C cal kit for the HP 8752C/53D/53E.

9-50 K ey De nitions

· · COLOR Color COLOR

Page 537 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 537 · click to enlarge
Page 537

T able 9-1.

Cross Reference of K ey Function to Programming Command (continued)

K ey Name Command

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CONVERSION [OFF] Conversion O CONV OFF

COPY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CORRECTION ON off Correction On CORRON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CORRECTION on OFF Correction O CORROFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

COUPLED CH ON off Coupled Channel On COUCON

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

COUPLED CH on OFF Coupled Channel O COUCOFF

NNNNNNNNNNNNNNNNNNNNNNN

CW FREQ CW Frequency CWFREQ

NNNNNNNNNNNNNNNNNNNNNNN

CW TIME CW Time CWTIME

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

D2/D1 to D2 ON off Ratio D2 to D1 On D1DIVD2ON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

D2/D1 to D2 on OFF Ratio D2 to D1 O D1DIVD2OFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DATA and MEMORY Data and Memory DISPD A TM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DATA ARRAY ON off Data Array On EXTMD A T A ON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DATA ARRAY on OFF Data Array O EXTMD A T A OFF

NNNNNNNNNNNNNNNNNNNNNNNNNN

DATA/MEM Ratio Data to Memory DISPDDM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DATA - MEM Data Minus Memory DISPDMM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DATA ! MEMORY Data to Memory D A TI

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DATA ONLY ON off Data Only On EXTMD A TOON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DATA ONLY on OFF Data Only O EXTMD A TOOFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DECR LOOP COUNTER Decrement Loop Counter DECRLOOC

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DEFAULT COLORS Default Colors DEFC

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DEFAULT PLOT SETUP Default Plot Setup DFL T

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DEFAULT PRINT SETUP Default Print Setup DEFLPRINT

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DEFINE STANDARD De ne Standard DEFS

NNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DELETE ALL FILES Delete All Files CLEARALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DELTA LIMITS Delta Limits LIMD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DEMODAMPLITUDE Demodulation Amplitude DEMO AMPL

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DEMODOFF Demodulation O DEMOOFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DEMODPHASE Demodulation Phase DEMOPHAS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DIRECTORY SIZE Directory Size DIRS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DISK UNIT NUMBER Disk Unit Number DISCUNIT

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DISP MKRS ON off Display Markers On DISM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DISP MKRS on OFF Display Markers O DISM

K ey De nitions 9-51

Show 5 more labels / short text on this page
CONTINUOUS Continuous CONT · · Copy MENUCOPY · DELAY Delay DELA · DELETE Delete SDEL
Page 538 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 538 · click to enlarge
Page 538

T able 9-1.

Cross Reference of K ey Function to Programming Command (continued)

K ey Name Command

DISPLA Y

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DISPLAYDATA Display Data DISPD A T A

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DO SEQUENCE Do Sequence DOSEn

NNNNNNNNNNNNNN

NNNNNNNNNNNNNN

DONE (Segment) Done SDON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DONE 1-PORT CAL Done 1-P ort Calibration SA V1

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DONE 2-PORT CAL Done 2-P ort Calibration SA V2

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DONE RESPONSE Done Response RESPDONE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DONE RESP ISOL'N CAL Done Response Isolation Cal RAID

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DONE SEQ MODIFY Done Sequence Modify DONM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DONE TRL/LRM Done TRL/LRM SA VT

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DOWN CONVERTER Down Converter DCONV

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DUAL CH ON off Dual Channel On DU A CON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DUAL CH on OFF Dual Channel O DU A COFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DUPLICATE SEQUENCE Duplicate Sequence DUPLSEQxSEQy

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EACH SWEEP Calibrate Each Sweep PWMCEA CS

NNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EDIT LIMIT LINE Edit Limit Line EDITLIML

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EDIT LIST Edit List EDITLIST

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ELECTRICAL DELAY Electrical Delay ELED

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EMIT BEEP Emit Beep EMIB

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

END SWEEP HIGH PULSE End Sweep High Pulse TTLHPULS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

END SWEEP LOW PULSE End Sweep Low Pulse TTLLPULS

Entry O ENTO

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EXT TRIG ON POINT External Trigger on P oint EXTTPOIN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EXT TRIG ON SWEEP External Trigger on Sweep EXTTON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EXTENSION INPUT A Extension Input A PORT A

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EXTENSION INPUT B Extension Input B PORTB

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EXTENSION PORT 1 Extension P ort 1 PORT1

9-52 K ey De nitions

Show 6 more labels / short text on this page
· Display MENUDISP · DONE Done EDITDONE · EDIT Edit SEDI · · Entry O
Page 539 of the Agilent Technologies 8753D user manual — SPECTTFM
Page 539 · click to enlarge
Page 539

T able 9-1.

Cross Reference of K ey Function to Programming Command (continued)

K ey Name Command

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EXTENSION PORT 2 Extension P ort 2 PORT2

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EXTENSIONS ON off Extensions On POREON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EXTENSIONS on OFF Extensions O POREOFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EXTERNAL DISK External Disk EXTD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FILENAME FILE0 File Name File 0 TITF0

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FILETITLE FILE0 File Name File 0 TITF0

N

NNNNNNNNNNNNNNNN

FIXED Fixed Load FIXE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FIXED MKR AUX VALUE Fixed Marker A uxiliary V alue MARKF A UV

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FIXED MKR POSITION Fixed Marker P osition DELRFIXM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FIXED MKR STIMULUS Fixed Marker Stimulus MARKFSTI

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FIXED MKR VALUE Fixed Marker V alue MARKFV AL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FLAT LINE Flat Line LIMTFL

F ORMA T

F ormat MENUFORM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FORMAT ARY ON off F ormat Array On EXTMFORMON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FORMAT ARY on OFF F ormat Array O EXTMFORMOFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FORMATDOS F ormat DOS FORMA TDOS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FORMATLIF F ormat LIF FORMA TLIF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FORMAT EXT DISK F ormat External Disk INIE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FORMAT INT DISK F ormat Internal Disk INID

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FORMAT INT MEMORY F ormat Internal Memory INTM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FREQ OFFS ON off Frequency O set On FREQOFFSON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FREQ OFFS on OFF Frequency O set O FREQOFFSOFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FREQUENCY BLANK Frequency Blank FREO

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FREQUENCYCW Frequency: CW LOFREQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FREQUENCYSWEEP Frequency: SWEEP

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FULL 2-PORT Full 2-P ort CALIFUL2

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FULL PAGE Full P age FULP

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FWD ISOL'N ISOL'N STD F orward Isolation FWDI

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNN

FWD MATCH (Label Class) Label F orward Match LABEFWDM

LABETTFM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FWD MATCH (Specify Class) Specify F orward Match SPECFWDM

SPECTTFM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FWD MATCH THRU F orward Match Thru FWDM

K ey De nitions 9-53

· FREQUENCY Frequency CALFFREQ

Page 540 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 540 · click to enlarge
Page 540

T able 9-1.

Cross Reference of K ey Function to Programming Command (continued)

K ey Name Command

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FWD TRANS (Label Class) Label F orward Transmission LABEFWDT

LABETTFT

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FWD TRANS (Specify Class) Specify F orward Transmission SPECFWDT

SPECTTFT

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FWD TRANS THRU F orward Transmission Thru FWDT

NNNNNNNNNNNNNNNNNNNNNNNNNN

G+jB MKR G+jB Marker Readout SMIMGB

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GATECENTER Gate Center GA TECENT

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GATESPAN Gate Span GA TESP AN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GATESTART Gate Start GA TEST AR

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GATESTOP Gate Stop GA TESTOP

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GATE ON off Gate On GA TEOON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GATE on OFF Gate O GA TEOOFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GATE SHAPE MAXIMUM Gate Shape Maximum GA TSMAXI

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GATE SHAPE MINIMUM Gate Shape Minimum GA TSMINI

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GATE SHAPE NORMAL Gate Shape Normal GA TSNORM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GATE SHAPE WIDE Gate Shape Wide GA TSWIDE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GOSUB SEQUENCE GOSUB Sequence GOSUBn

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GRAPHICS ON off Graphics On EXTMGRAPON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GRAPHICS on OFF Graphics O EXTMGRAPOFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GRATICULE [ ] Print Color - graticule PCOLGRA T

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GRATICULE TEXT Graticule COLOGRA T

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

HARMONIC OFF Harmonic Mode O HARMOFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

HARMONIC SECOND Measure Second Harmonic HARMSEC

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

HARMONIC THIRD Measure Third Harmonic HARMTHIR

NNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

HP-IB DIAG ON off HP-IB Diagnostics On DEBUON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

HP-IB DIAG on OFF HP-IB Diagnostics O DEBUOFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

IF BW [ ] IF Bandwidth IFBW

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

IF LIMIT TEST FAIL If Limit T est F ail IFL TF AIL

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

IF LIMIT TEST PASS If Limit T est P ass IFL TP ASS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

IF LOOP COUNTER = 0 IF Loop Counter = 0 IFLCEQZE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

IF LOOP < > COUNTER 0 IF Loop < > Counter 0 IFLCNEZE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

IMAGINARY Imaginary IMA G

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INCR LOOP COUNTER Increment Loop Counter INCRLOOC

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

9-54 K ey De nitions

HOLD Hold HOLD · INTENSITY Intensity INTE

Page 541 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 541 · click to enlarge
Page 541

T able 9-1.

Cross Reference of K ey Function to Programming Command (continued)

K ey Name Command

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INTERNAL DISK Internal Disk INTD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INTERNAL MEMORY Select Internal Memory INTM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INTERPOL ON off Interpolation On CORION

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INTERPOL on OFF Interpolation O CORIOFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ISOLATION (2-P ort) Isolation ISOL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ISOLATION (One-P ath 2-P ort) Isolation ISOOP

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ISOLATION DONE Isolation Done ISOD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ISOL'N STD Isolation Standard RAIISOL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

KIT DONE (MODIFIED) Kit Done KITD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LABEL KIT Label Kit LABK

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LABEL STD Label Standard LABS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LEFT LOWER Left lower LEFL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LEFT UPPER Left Upper LEFU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LIMIT LINE ON off Limit Line On LIMILINEON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LIMIT LINE on OFF Limit Line O LIMILINEOFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LIMIT TEST ON off Limit T est On LIMITESTON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LIMIT TEST on OFF Limit T est O LIMITESTOFF

NNNNNNNNNNNNNNNNNNNNNNNNNN

LIN FREQ Linear Frequency LINFREQ

NNNNNNNNNNNNNNNNNNNNNNN

LIN MAG Linear Magnitude LINM

NNNNNNNNNNNNNNNNNNNNNNN

LIN MKR Linear Marker POLMLIN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LINE TYPE DATA Line Type Data LINTD A T A

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LINE TYPE MEMORY Line Type Memory LINTMEMO

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LIST FREQ List Frequency LISTFREQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LIST VALUES List V alues LISV

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LN/MATCH 1 Line/Match 1 TRLL1

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LN/MATCH 2 Line/Match 1 TRLL2

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LO CONTROL ON off LO Control On LOCONTON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LO CONTROL on OFF LO Control O LOCONTOFF

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LO SOURCE ADDRESS LO Source A ddress ADDRLSRC

NNNNNNNNNNNNNN

LOAD Load STDTLO AD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOAD NO OFFSET Load No O set LO AN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOAD OFFSET Load O set LO A O

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOAD SEQ FROM DISK Load Sequence From Disk LO ADSEQn

K ey De nitions 9-55

LINE/MATCH Line/Match

Page 542 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 542 · click to enlarge
Page 542

T able 9-1.

Cross Reference of K ey Function to Programming Command (continued)

K ey Name Command

LOCAL

NNNNNNNNNNNNNNNNNNNNNNNNNN

LOG FREQ Logarithmic Frequency LOGFREQ

NNNNNNNNNNNNNNNNNNNNNNN

LOG MAG Logarithmic Magnitude LOGM

NNNNNNNNNNNNNNNNNNNNNNN

LOG MKR Logarithmic Marker SMIMLOG

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOOP COUNTER Loop Counter LOOC

NNNNNNNNNNNNNN

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOW PASS IMPULSE Low P ass Impulse LOWPIMPU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOW PASS STEP Low P ass Step LOWPSTEP

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOWER LIMIT Lower Limit LIML

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MANUAL TRG ON POINT Manual Trigger On P oint MANTRIG

MARKER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER ! CENTER Marker to Center MARKCENT

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER ! CW Marker to CW MARKCW

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER ! DELAY Marker to Delay MARKDELA

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER ! MIDDLE Marker to Middle MARKMIDD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER ! REFERENCE Marker to Reference MARKREF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER ! SPAN Marker to Span MARKSP AN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER ! START Marker to Start MARKST AR

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER ! STIMULUS Marker to Stimulus MARKSTIM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER ! STOP Marker to Stop MARKSTOP

NNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER 1 Marker 1 MARK1

NNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER 2 Marker 2 MARK2

NNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER 3 Marker 3 MARK3

NNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER 4 Marker 4 MARK4

NNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER 5 Marker 5 MARK5

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER all OFF All Markers O MARKOFF

MARKER F CTN

Marker Function MENUMRKF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKERSCONTINUOUS Markers Continuous MARKCONT

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKERSCOUPLED Markers Coupled MARKCOUP

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKERSDISCRETE Markers Discrete MARKDISC

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKERSUNCOUPLED Markers Uncoupled MARKUNCO

9-56 K ey De nitions

Show 6 more labels / short text on this page
· Local · LOSS Loss POWLLOSS · · Marker MENUMARK ·
Page 543 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 543 · click to enlarge
Page 543

T able 9-1.

Cross Reference of K ey Function to Programming Command (continued)

K ey Name Command

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MAXIMUM FREQUENCY Maximum Frequency MAXF

MEAS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MEASURE RESTART Measure Restart REST

NNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MIDDLE VALUE Middle V alue LIMM

NNNNNNNNNNNNNNNNNNNNNNN

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MINIMUM FREQUENCY Minimum Frequency MINF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MKR SEARCH [OFF] Marker Search O SEA OFF

NNNNNNNNNNNNNNNNNNNNNNNNNN

MKR ZERO Marker Zero MARKZERO

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MODIFY [ ] Modify Kit MODI1

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NETWORK ANALYZER Network Analyzer INSMNET A

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NEW SEQ/MODIFY SEQ New Sequence/Modify Sequence NEWSEn

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NEXT PAGE Display Next P age of T abular NEXP

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NUMBER OF GROUPS Number of Groups NUMG

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NUMBER OF POINTS Number of P oints POIN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NUMBER OF READINGS Number of Readings NUMR

NNNNNNNNNNN

OFF O CONOFF

NNNNNNNNNNNNNNNNNNNNNNN

OFFSET O set OFLS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

OFFSET DELAY O set Delay OFSD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

OFFSET LOADS DONE O set Loads Done OFLD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

OFFSET LOSS O set Loss OFSL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

OFFSET Z0 O set Impedance OFSZ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

OMIT ISOLATION Omit Isolation OMII

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ONE-PATH 2-PORT One-P ath 2-P ort CALIONE2

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ONE SWEEP Calibrate One Sweep PWMCONES

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

OP PARMS MKRS etc T abular Listing of Operating OPEP

K ey De nitions 9-57

Show 6 more labels / short text on this page
· Measure MENUMEAS · MEMORY Memory DISPMEMO · MINIMUM Minimum WINDMINI · Listing · P arameters
Page 544 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 544 · click to enlarge
Page 544

T able 9-1.

Cross Reference of K ey Function to Programming Command (continued)

K ey Name Command

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

P MTR/HPIB TO TITLE P ower Meter HPIB to Title PMTRTTIT

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PARALL IN BIT NUMBER P arallel in Bit Number P ARAIN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PARALL IN IF BIT H P arallel in IF Bit H IFBIHIGH

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PARALL IN IF BIT L P arallel in IF Bit L IFBILOW

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PARALLEL [COPY] Set parallel port to copy mode P ARALCPY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PARALLEL [GPIO] Set parallel port to GPIO mode P ARALGPIO

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PARALLEL OUT ALL P arallel Out All P ARA OUT

NNNNNNNNNNNNNNNNN

PAUSE P ause P A US

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PAUSE TO SELECT P ause to Select PTOS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PEN NUM DATA P en Number Data PENND A T A

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PEN NUM GRATICULE P en Number Graticule PENNGRA T

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PEN NUM MARKER P en Number Marker PENNMARK

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PEN NUM MEMORY P en Number Memory PENNMEMO

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PEN NUM TEXT P en Number T ext PENNTEXT

NNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PHASE OFFSET Phase O set PHA O

NNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOT DATA ON off Plot Data On PD A T A ON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOT DATA on OFF Plot Data O PD A T A OFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOT GRAT ON off Plot Graticule On PGRA TON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOT GRAT on OFF Plot Graticule O PGRA TOFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOT MEM ON off Plot Memory On PMEMON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOT MEM on OFF Plot Memory O PMEMOFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOT MKR ON off Plot Marker ON PMKRON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOT MKR on OFF Plot Marker O PMKROFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOTNAME PLOTFILE Plot name Plot le TITP

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOT SPEED [FAST] Plot Speed F ast PLOSF AST

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOT SPEED [SLOW] Plot Speed Slow PLOSSLOW

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOT TEXT ON off Plot T ext On PTEXTON

9-58 K ey De nitions

PHASE Phase PHAS · PLOT Plot PLOT

Page 545 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 545 · click to enlarge
Page 545

T able 9-1.

Cross Reference of K ey Function to Programming Command (continued)

K ey Name Command

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOT TEXT on OFF Plot T ext O PTEXTOFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOTTER BAUD RATE Plotter Baud Rate PL TTRB A UD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOTTER FORM FEED Plotter F orm F eed PL TTRFORF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLTR PORTDISK Plotter P ort Disk PL TPRTDISK

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLTR PORTHPIB Plotter P ort HPIB PL TPRTHPIB

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLTR PORTPARALLEL Plotter P ort P arallel PL TPRTP ARA

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLTR PORTSERIAL Plotter P ort Serial PL TPRTSERI

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLTR TYPE [PLOTTER] Plot to a Plotter PL TTYPPL TR

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLTR TYPE [HPGL PRT] Plot to a HP-GL/2 Compatible PL TTYPHPGL

NNNNNNNNNNNNNNNNN

POLAR P olar POLA

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PORT PWR [COUPLED] P ort P ower Coupled PORTPCPLD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PORT PWR [UNCOUPLED] P ort P ower Uncoupled PORTPUNCPLD

NNNNNNNNNNNNNNNNN

POWER P ower POWE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

POWERFIXED P ower Fixed LOPOWER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

POWERSWEEP P ower Sweep Mode LOPSWE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

POWER MTR P ower Meter POWM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

POWER MTR[436A] P ower Meter 436A POWMON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

POWER MTR[437B/438A] P ower Meter 437B/438A POWMOFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

POWER RANGES P ower Ranges PWRR

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

POWER SWEEP P ower Sweep POWS

PRESET

F actory Preset RST

PRES

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRESETFACTORY F actory Preset RST

PRES

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PREVIOUS PAGE Previous P age PREP

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRINT ALL COLOR Print Entire List - Color PRINT ALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRINT ALL MONOCHROME Print Entire List - Monochrome PRINT ALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRINTCOLOR Selects Color Printer PRIC

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRINT COLOR Print Color PRINALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRINTMONOCHROME Selects Monochrome Printer PRIS

K ey De nitions 9-59

Printer ·

Page 546 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 546 · click to enlarge
Page 546

T able 9-1.

Cross Reference of K ey Function to Programming Command (continued)

K ey Name Command

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRINT MONOCHROME Print Monochrome PRINALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRINT SEQUENCE Print Sequence PRINSEQn

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRINTER BAUD RATE Printer Baud Rate PRNTRB A UD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRINTER FORM FEED Printer F orm F eed PRNTRFORF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRNTR PORTHPIB Printer P ort HPIB PRNPRTHPIB

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRNTR PORTPARALLEL Printer P ort P arallel PRNPRTP ARA

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRNTR PORTSERIAL Printer P ort Serial PRNPRTSERI

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRNTR TYPE [DESKJET] DeskJet Printer PRNTYPDJ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRNTR TYPE [EPSON-P2] EPSON ESC/P2 Printer Central PRNTYPEP

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRNTR TYPE [LASERJET] LaserJet Printer PRNTYPLJ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRNTR TYPE [PAINTJET] P aintJet Printer PRNTYPP J

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRNTR TYPE [THINKJET] ThinkJet Printer PRNTYPT J

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PWR LOSS ON off P ower Loss On PWRLOSSON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PWR LOSS on OFF P ower Loss O PWRLOSSOFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PWR RANGE AUTO man P ower Range A uto PWRRP A UTO

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PWR RANGE auto MAN P ower Range Man PWRRPMAN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PWRMTR CAL [ ] P ower Meter Calibration CALPOW

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PWRMTR CAL [OFF] P ower Meter Calibration O PWRMMCALOFF

NNNNN

R Measure Channel R MEASR

NNNNNNNNNNNNNNNNNNNNNNNNNN

R+jX MKR R+jX Marker Readout SMIMRX

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RANGE 0 -15 TO +10 P ower Range 0 PRAN0

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RANGE 1 -25 TO 0 P ower Range 1 PRAN1

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RANGE 2 -35 TO -10 P ower Range 2 PRAN2

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RANGE 3 -45 TO -20 P ower Range 3 PRAN3

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RANGE 4 -55 TO -30 P ower Range 4 PRAN4

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RANGE 5 -65 TO -40 P ower Range 5 PRAN5

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RANGE 6 -75 TO -50 P ower Range 6 PRAN6

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RANGE 7 -85 TO -60 P ower Range 7 PRAN7

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RAW ARRAY ON oFF Raw Array On EXTMRA WON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RAW ARRAY on OFF Raw Array O EXTMRA WOFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RAW OFFSET ON Off Raw O set RA WOFFON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RAW OFFSET On OFF Raw O set RA WOFFSOFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Re/Im MKR Real/Imaginary Markers POLMRI

9-60 K ey De nitions

Language

Page 547 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 547 · click to enlarge
Page 547

T able 9-1.

Cross Reference of K ey Function to Programming Command (continued)

K ey Name Command

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

READ FILE TITLES Read File Titles REFT

NNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECALL COLORS Recall Colors RECO

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECALL REG1 Recall Register 1 RECA1

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECALL REG2 Recall Register 2 RECA2

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECALL REG3 Recall Register 3 RECA3

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECALL REG4 Recall Register 4 RECA4

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECALL REG5 Recall Register 5 RECA5

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECALL REG6 Recall Register 6 RECA6

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECALL REG7 Recall Register 7 RECA7

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECALL STATE Recall State RECA

RECAREG

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECEIVER CAL Receiver Calibration REIC

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REFERENCE POSITION Reference P osition REFP

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REFERENCE VALUE Reference V alue REFV

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REFLFWD S11 (A/R) Re ection F orward S11 A/R RFLP

S11

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REFLREV S22 (B/R) Re ection Reverse S22 B/R S22

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REFLECTION Re ection REFOP

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REMOVE ADAPTER Remove A dapter MODS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RESET COLOR Reset Color RSCO

NNNNNNNNNNNNNNNNNNNNNNNNNN

RESPONSE (Calibrate) Response CALIRESP

NNNNNNNNNNNNNNNNNNNNNNNNNN

RESPONSE (Label Class) Response LABERESP

NNNNNNNNNNNNNNNNNNNNNNNNNN

RESPONSE (Specify Class) Response SPECRESP

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RESPONSE & ISOL'N Response and Isolation CALIRAI

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RESPONSE & ISOL'N Response and Isolation LABERESI

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RESPONSE & ISOL'N Response and Isolation SPECRESI

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RESTORE DISPLAY Restore Display RESD

K ey De nitions 9-61

Show 4 more labels / short text on this page
REAL Real REAL · (Calibrate) · (Label Class) · (Specify Class)
Page 548 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNN
Page 548 · click to enlarge
Page 548

T able 9-1.

Cross Reference of K ey Function to Programming Command (continued)

K ey Name Command

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RESUME CAL SEQUENCE Resume Calibration Sequence RESC

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REV ISOL'N ISOL'N STD Reverse Isolation REVI

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REV MATCH (Label Class) Label Reverse Match LABEREVM

LABETTRM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REV MATCH (Specify Class) Specify Reverse Match SPECREVM

SPECTTRM

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REV MATCH THRU Reverse Match Thru REVM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REV TRANS (Label Class) Label Reverse Transmission LABEREVT

LABETTRT

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REV TRANS (Specify Class) Specify Reverse Transmission SPECREVT

SPECTTRT

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REV TRANS THRU Reverse Transmission Thru REVT

NNNNNNNNNNNNNNNNNNNNNNN

RF > LO RF Greater Than LO RFGTLO

NNNNNNNNNNNNNNNNNNNNNNN

RF < LO RF Less Than LO RFL TLO

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RIGHT LOWER Right Lower RIGL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RIGHT UPPER Right Upper RIGU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

S11 1-PORT S

NNNNNNNNNNNNNN

S11A S

A Re ected F orward Match LABES11A

NNNNNNNNNNNNNN

S11A S

A Re ected F orward Match SPECS11A

NNNNNNNNNNNNNN

S11B S

B Line F orward Match LABES11B

NNNNNNNNNNNNNN

S11B S

B Line F orward Match SPECS11B

NNNNNNNNNNNNNN

S11C S

C Line F orward Transmission LABES11C

NNNNNNNNNNNNNN

S11C S

C Line F orward Transmission SPECS11C

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

S11 REFL OPEN S

Re ect Short TRLR1

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

S22 1-PORT S

NNNNNNNNNNNNNN

S22A S

A Re ected Reverse Match LABES22A

NNNNNNNNNNNNNN

S22A S

A Re ected Reverse Match SPECS22A

9-62 K ey De nitions

Show 10 more labels / short text on this page
1-P ort CALIS111 · (Label Class) LABETRFM · (Specify Class) SPECTRFM · (Label Class) LABETRRM · (Specify Class) SPECTRRM · (Label Class) LABETLFT · (Specify Class) SPECTLFT · 1-P ort CALIS221 · (Label Class) LABETRRM · (Specify Class) SPECTRRM
Page 549 of the Agilent Technologies 8753D user manual — Q3
Page 549 · click to enlarge
Page 549

T able 9-1.

Cross Reference of K ey Function to Programming Command (continued)

K ey Name Command

NNNNNNNNNNNNNN

S22B S

B Line Reverse Transmission LABES22B

NNNNNNNNNNNNNN

S22B S

B Line Reverse Transmission SPECS22B

NNNNNNNNNNNNNN

S22C S

C Line Reverse Transmission LABES22C

N

NNNNNNNNNNNNN

S22C S

C Line Reverse Transmission SPECS22C

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

S22 REFL OPEN S

Re ect Short TRLR2

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SAMPLR COR ON off Sampler Correction On SAMCON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SAMPLR COR on OFF Sampler Correction O SAMCOFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SAVE COLORS Save Colors SV CO

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SAVE USER KIT Save User Kit SA VEUSEK

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SAVE USING ASCII Save ASCII F ormat SA VU ASCI

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SAVE USING BINARY Save Using Binary SA VUBINA

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SCALE PLOT [FULL] Scale Plot Full SCAPFULL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SCALE PLOT [GRAT] Scale Plot Graticule SCAPGRA T

Scale Reference MENUSCAL

SCALE REF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEARCH LEFT Search Left SEAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEARCH RIGHT Search Right SEAR

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEARCHMAX Search Maximum SEAMAX

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEARCHMIN Search Minimum SEAMIN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEARCHOFF Search O SEA OFF

NNNNNNNNNNNNNNNNNNNN

SECOND Second Harmonic HARMSEC

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEGMENTCENTER Segment Center CENT

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEGMENTSPAN Segment Span SP AN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEGMENTSTART Segment Start ST AR

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEGMENTSTOP Segment Stop STOP

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQUENCE 1 SEQ1 Select Sequence 1 SEQ1

Q1

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQUENCE 2 SEQ2 Select Sequence 2 SEQ2

Q2

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQUENCE 3 SEQ3 Select Sequence 3 SEQ3

Q3

K ey De nitions 9-63

Show 6 more labels / short text on this page
(Label Class) LABETLRM · (Specify Class) SPECTLRM · (Label Class) LABETLRT · (Specify Class) SPECTLRT · SCALE/DIV Scale/Division SCAL ·
Page 550 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 550 · click to enlarge
Page 550

T able 9-1.

Cross Reference of K ey Function to Programming Command (continued)

K ey Name Command

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQUENCE 4 SEQ4 Select Sequence 4 SEQ4

Q4

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQUENCE 5 SEQ5 Select Sequence 5 SEQ5

Q5

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQUENCE 6 SEQ6 Select Sequence 6 SEQ6

Q6

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQUENCE 1 SEQ1 Select Sequence 1 to Title TITSEQ1

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQUENCE 2 SEQ2 Select Sequence 2 to Title TITSEQ2

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQUENCE 3 SEQ3 Select Sequence 3 to Title TITSEQ3

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQUENCE 4 SEQ4 Select Sequence 4 to Title TITSEQ4

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQUENCE 5 SEQ5 Select Sequence 5 to Title TITSEQ5

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQUENCE 6 SEQ6 Select Sequence 6 to Title TITSEQ6

NNNNNNNNNNNNNNNNNNNNNNN

SET BIT Set Bit SETBIT

NNNNNNNNNNNNNNNNNNNNNNNNNN

SET DATE Set Date SETD A TE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SET FREQ LOW PASS Set Frequency Low P ass SETF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SET REFREFLECT Set Reference: Re ect SETRREFL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SET REFTHRU Set Reference: Thru SETRTHRU

NNNNNNNNNNNNNNNNNNNNNNNNNN

SET TIME Set Time SETTIME

NNNNNNNNNNNNNNNNNNNN

SET Z0 Set Impedance SETZ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SHOW MENUS Show Menus SHOM

NNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SINGLE POINT Single P oint LIMTSP

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SINGLE SEG SWEEP Single Segment Sweep SSEG

NNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SLOPE ON off Slope On SLOPON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SLOPE on OFF Slope On SLOPOFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SLOPING LINE Sloping Line LIMTSL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SMITH CHART Smith Chart SMIC

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SMOOTHING APERTURE Smoothing Aperture SMOO APER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SMOOTHING ON off Smoothing On SMOOON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SMOOTHING on OFF Smoothing O SMOOOFF

9-64 K ey De nitions

SINGLE Single SING · SLIDING Sliding SLIL · SLOPE Slope SLOPE

Page 551 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 551 · click to enlarge
Page 551

T able 9-1.

Cross Reference of K ey Function to Programming Command (continued)

K ey Name Command

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SOURCE PWR ON off Source P ower On SOUPON

POWTOFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SOURCE PWR on OFF Source P ower O SOUPOFF

POWTON

NNNNNNNNNNNNNN

SPAN Span SP AN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SPECIFY GATE Specify Gate SPEG

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SPLIT DISP ON off Split Display On SPLDON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SPLIT DISP on OFF Split Display O SPLDOFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SPUR AVOID ON Off Spur A voidance On SM8ON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SPUR AVOID On OFF Spur A voidance O SM8OFF

ST ART

Start LOFST AR

NNNNNNNNNNNNNNNNN

START Start ST AR

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STATS ON off Statistics On MEAST A TON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STATS on OFF Statistics O MEAST A TOFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STD DONE (DEFINED) Standard Done STDD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STD TYPEARBITRARY Standard Type: Arbitrary Impedance STDT ARBI

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

IMPEDANCE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STD TYPEDELAY/THRU Standard Type: Delay/Thru STDTDELA

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STD TYPELOAD Standard Type: Load STDTLO AD

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STD TYPEOPEN Standard Type: Open STDTOPEN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STD TYPESHORT Standard Type: Short STDTSHOR

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STEP SIZE Step Size STPSIZE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STIMULUS VALUE Stimulus V alue LIMS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STIMULUS OFFSET Stimulus O set LIMISTIO

STOP

NNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STORE SEQ TO DISK Store Sequence to Disk STORSEQn

NNNNNNNNNNNNNNNNN

SWEEP Sweep Mode LOFSWE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SWEEP TIME [ ] Sweep Time SWET

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SWEEP TIME [AUTO] Sweep Time SWEA

NNNNNNNNNNN

SWR SWR SWR

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SYSTEM CONTROLLER System Controller

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TAKE CAL SWEEP T ake Calibration Sweep T AKCS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TAKE RCVR CAL SWEEP T ake Receiver Calibration Sweep T AKRS

K ey De nitions 9-65

Show 6 more labels / short text on this page
· · Stop LOFSTOP · STOP Stop STOP · · System MENUSYST
Page 552 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 552 · click to enlarge
Page 552

T able 9-1.

Cross Reference of K ey Function to Programming Command (continued)

K ey Name Command

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TALKER/LISTENER T alker/Listener T ALKLIST

NNNNNNNNNNNNNNNNNNNN

TARGET T arget SEA T ARG

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TERMINAL IMPEDANCE T erminal Impedance TERI

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TESTPORT (1) 2 T estport 1 TSTPP1

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TESTPORT 1 (2) T estport 2 TSTPP2

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TESTSET I/O FWD T estset I/O F orward TSTIOFWD

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TESTSET I/O REV T estset I/O Reverse TSTIOREV

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TESTSET SW XXXX T estset Switching XXXX TSSWI

CSWI

NNNNNNNNNNNNNN

TEXT T ext COLOTEXT

NNNNNNNNNNNNNNNNNNNNNNNNNN

TEXT [ ] Print Color - T ext PCOL TEXT

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

THRU THRU Thru Thru TRL T

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TIME STAMP ON off Time Stamp On TIMEST AMON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TIME STAMP on OFF Time Stamp O TIMEST AMOFF

NNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TITLE FILE1 Title File 1 TITF1

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TITLE FILE2 Title File 2 TITF2

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TITLE FILE3 Title File 3 TITF3

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TITLE FILE4 Title File 4 TITF4

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TITLE FILE5 Title File 5 TITF5

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TITLE SEQUENCE Title Sequence TITSQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TITLE TO MEMORY Title to Memory TITTMEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TITLE TO P MTR/HPIB Title to P ower Meter/HPIB TITTPMTR

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TITLE TO PERIPHERAL Title to HP-IB P eripheral TITTPERI

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TITLE TO PRNTR/HPIB Title to HP-IB Printer TITTPRIN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRACKING ON off Tracking On TRA CKON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRACKING on OFF Tracking O TRA CKOFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRANS DONE Transmission Done TRAD

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRANSFWD S21 (B/R) Transmission F orward S21 B/R S21

TRAP

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRANSREV S12 (A/R) Transmission Reverse S12 A/R S12

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRANSFORM ON off Transform On TIMDTRANON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRANSFORM on OFF Transform O TIMDTRANOFF

9-66 K ey De nitions

TINT Tint TINT · TITLE Title TITL

Page 553 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNN
Page 553 · click to enlarge
Page 553

T able 9-1.

Cross Reference of K ey Function to Programming Command (continued)

K ey Name Command

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRIGGERTRIG OFF External Trigger O EXTTOFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRL*/LRM* 2-PORT Thru, Re ect, Line/Line , Re ect,

CALITRL2

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRL/LRM OPTION Thru, Re ect, Line/Line , Re ect,

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRL LINE OR MATCH TRL Line or Match SPECTRLL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRL LINE OR MATCH TRL Line or Match LABETRLL

NNNNNNNNNNNNNNNNNNNNNNNNNN

TRL THRU TRL Thru SPECTRL T

NNNNNNNNNNNNNNNNNNNNNNNNNN

TRL THRU TRL Thru LABETRL T

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRL REFLECT TRL Re ect SPECTRLR

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRL REFLECT TRL Re ect LABETRLR

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TTL OUT HIGH TTL Out High TTLOH

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TTL OUT LOW TTL Out Low TTLOL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TUNED RECEIVER Tuned Receiver INSMTUNR

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

UP CONVERTER Up Converter UCONV

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

UPPER LIMIT Upper Limit LIMU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

USE MEMORY ON off Use Memory On WINDUSEMON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

USE MEMORY on OFF Use Memory O WINDUSEMOFF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

USE PASS CONTROL Use P ass Control USEP ASC

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

USE SENSOR (A) / B Use Sensor A ENSA

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

USE SENSOR A / (B) Use Sensor B ENSB

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

VELOCITY FACTOR V elocity F actor VELOF A CT

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

VIEW MEASURE View Measure VIEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

VOLUME NUMBER V olume Number DISCV OLU

NNNNNNNNNNNNNNNNNNNN

WAIT x W ait x Seconds SEQW AIT

K ey De nitions 9-67

Show 10 more labels / short text on this page
TRANSMISSION Transmission FWDT · Match · Match · (Specify Class) · (Label Class) · (Specify Class) · (Label Class) · (Specify Class) · (Label Class) · UNCOUPLED Uncoupled UNCPLD
Page 554 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNN
Page 554 · click to enlarge
Page 554

T able 9-1.

Cross Reference of K ey Function to Programming Command (continued)

K ey Name Command

NNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

WAVEGUIDE W aveguide W A VE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

WAVEGUIDE DELAY W aveguide Delay W A VD

NNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

WIDTH VALUE Width V alue WIDV

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

WIDTHS ON off Widths On WIDTON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

WIDTHS on OFF Widths O WIDTOFF

NNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

WINDOWMAXIMUM Window Maximum WINDMAXI

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

WINDOWMINIMUM Window Minimum WINDMINI

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

WINDOWNORMAL Window Normal WINDNORM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

XMIT CNTRL [Xon-Xoff] Transmit Control (printer) PRNHNDSHKXON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

XMIT CNTRL [DTR-DSR] Transmit Control (printer) PRNHNDSHKDTR

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

XMIT CNTRL [Xon-Xoff] Transmit Control (plotter) PL THNDSHKXON

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

XMIT CNTRL [DTR-DSR] Transmit Control (plotter) PL THNDSHKDTR

NNNNNNNNNNNNNNNNNNNNNNN

Y: REFL Y : Re ection CONVYREF

NNNNNNNNNNNNNNNNNNNNNNNNNN

Y: TRANS Y : Transmission CONVYTRA

NNNNNNNNNNNNNNNNNNNN

YELLOW Y ellow YELLOW

NNNNNNNNNNNNNNNNNNNNNNN

Z: REFL Z: Re ection CONVZREF

NNNNNNNNNNNNNNNNNNNNNNNNNN

Z: TRANS Z: Transmission CONVZTRA

9-68 K ey De nitions

WHITE White WHITE · WINDOW Window WINDOW

Page 555 of the Agilent Technologies 8753D user manual — Chapter 9 Key Definitions
Page 555 · click to enlarge
Page 555

The following table lists the softkey functions alphabetically , and the corresponding front-panel

access key . This table is useful in determining which front-panel key leads to a speci c softkey .

K ey De nitions 9-69

Softkey Locations

Page 556 of the Agilent Technologies 8753D user manual — MEAS
Page 556 · click to enlarge
Page 556

T able 9-2. Softkey Locations

A ccess K ey

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MODE MENU

MARKER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MODE OFF

MARKER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REF = 1

MARKER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REF = 2

MARKER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REF = 3

MARKER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REF = 4

MARKER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REF = 5

MARKER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REF = FIXED MKR

MARKER

NNNNNNNNNNN

1/S

MEAS

NNNNN

A

MEAS

NNNNNNNNNNN

A/B

MEAS

NNNNNNNNNNN

A/R

MEAS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ACTIVE ENTRY

DISPLA Y

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ACTIVE MRK MAGNITUDE

DISPLA Y

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ADAPTER:COAX

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ADAPTER:WAVEGUIDE

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ADAPTER DELAY

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ADAPTER REMOVAL

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ADDRESS: 8753

LOCAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ADDRESS: CONTROLLER

LOCAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ADDRESS: DISK

LOCAL

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ADDRESS: DISK

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ADDRESS: P MTR/HPIB

LOCAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ADJUST DISPLAY

DISPLA Y

NNNNNNNNNNNNNNNNNNNNNNN

ALL OFF

MARKER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ALL SEGS SWEEP

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ALTERNATE A and B

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AMPLITUDE

SYSTEM

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AMPLITUDE OFFSET

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ANALOG IN Aux Input

MEAS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ARBITRARY IMPEDANCE

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ASSERT SRQ

SEQ

9-70 K ey De nitions

Show 33 more labels / short text on this page
Softkey Front-P anel · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ·
Page 557 of the Agilent Technologies 8753D user manual — CAL
Page 557 · click to enlarge
Page 557

T able 9-2. Softkey Locations (continued)

A ccess K ey

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AUTO FEED on OFF

COPY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AUTO SCALE

SCALE REF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AVERAGING FACTOR

A VG

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AVERAGING on OFF

A VG

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AVERAGING RESTART

A VG

NNNNN

B

MEAS

N

NNNNNNNNNN

B/R

MEAS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

BACKGROUND INTENSITY

DISPLA Y

NNNNNNNNNNNNNNNNNNNNNNNNNN

BANDPASS

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

BEEP DONE ON off

DISPLA Y

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

BEEP FAIL on OFF

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

BEEP WARN on OFF

DISPLA Y

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

BLANK DISPLAY

DISPLA Y

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

BRIGHTNESS

DISPLA Y

NNNNNNNN

C0

CAL

NNNNNNNN

C1

CAL

NNNNNNNN

C2

CAL

NNNNNNNN

C3

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL FACTOR

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL FACTOR SENSOR A

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL FACTOR SENSOR B

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL KIT [ ]

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL KIT2.4mm

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL KIT: 2.92*

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL KIT2.92mm

CAL

K ey De nitions 9-71

Show 26 more labels / short text on this page
Softkey Front-P anel · · · · · · · · · · · · · · · · · · · · · · · · ·
Page 558 of the Agilent Technologies 8753D user manual — CAL
Page 558 · click to enlarge
Page 558

T able 9-2. Softkey Locations (continued)

A ccess K ey

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL KIT3.5mmC

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL KIT3.5mmD

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL KITTRL 3.5mm

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL KIT7mm

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL KIT: N 50

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL KIT: N 75

CAL

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL KIT: USER KIT

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL ZO: LINE ZO

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CAL ZO: SYSTEM ZO

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CALIBRATE MENU

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CALIBRATE: NONE

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CH1 DATA [ ]

COPY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CH1 DATA LIMIT LN

DISPLA Y

NNNNNNNNNNNNNNNNNNNNNNN

CH1 MEM

DISPLA Y

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CH1 MEM [ ]

COPY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CH2 DATA [ ]

COPY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CH2 DATA LIMIT LN

DISPLA Y

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CH2 MEM [ ]

COPY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CH2 MEM REF LINE

DISPLA Y

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CH PWR [COUPLED]

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CH PWR [UNCOUPLED]

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CHOP A and B

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CLEAR BIT

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CLEAR LIST

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CLEAR SEQUENCE

SEQ

NNNNNNNNNNNNNN

COAX

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

COAXIAL DELAY

SCALE REF

NNNNNNNNNNNNNNNNN

COLOR

DISPLA Y

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CONFIGURE

SYSTEM

N

CONFIGURE EXTERNAL DISK

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CONTINUE SEQUENCE

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CONTINUOUS

MENU

9-72 K ey De nitions

Show 33 more labels / short text on this page
Softkey Front-P anel · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ·
Page 559 of the Agilent Technologies 8753D user manual — SA VE/RECALL
Page 559 · click to enlarge
Page 559

T able 9-2. Softkey Locations (continued)

A ccess K ey

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CONVERSION [ ]

MEAS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

CORRECTION on OFF

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

COUPLED CH on OFF

MENU

NNNNNNNNNNNNNNNNNNNNNNN

CW FREQ

MENU

NNNNNNNNNNNNNNNNNNNNNNN

CW TIME

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

D2/D1 to D2 on OFF

DISPLA Y

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DATA and MEMORY

DISPLA Y

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DATA ARRAY on OFF

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNN

DATA/MEM

DISPLA Y

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DATA - MEM

DISPLA Y

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DATA ! MEMORY

DISPLA Y

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DATA ONLY on OFF

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DECISION MAKING

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DECR LOOP COUNTER

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DEFAULT COLORS

DISPLA Y

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DEFAULT PLOT SETUP

COPY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DEFAULT PRINT SETUP

COPY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DEFINE DISK-SAVE

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DEFINE PLOT

COPY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DEFINE PRINT

COPY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DEFINE STANDARD

CAL

NNNNNNNNNNNNNNNNN

DELAY

F ORMA T

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DELAY/THRU

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DELETE ALL FILES

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DELETE FILE

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DELTA LIMITS

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DEMOD: AMPLITUDE

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DEMOD: OFF

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DEMOD: PHASE

SYSTEM

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DIRECTORY SIZE (LIF)

SA VE/RECALL

K ey De nitions 9-73

Show 31 more labels / short text on this page
Softkey Front-P anel · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ·
Page 560 of the Agilent Technologies 8753D user manual — SYSTEM
Page 560 · click to enlarge
Page 560

T able 9-2. Softkey Locations (continued)

A ccess K ey

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DISK UNIT NUMBER

LOCAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DISK UNIT NUMBER

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DISPLAY: DATA

DISPLA Y

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DISP MKRS ON off

MARKER F CTN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DO BOTH FWD + REV

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DO SEQUENCE

SEQ

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DONE 1-PORT CAL

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DONE 2-PORT CAL

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DONE RESPONSE

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DONE RESP ISOL'N CAL

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DONE SEQ MODIFY

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DONE TRL/LRM

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DOWN CONVERTER

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DUAL CH on OFF

DISPLA Y

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DUMP GRAPH on OFF

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DUPLICATE SEQUENCE

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EACH SWEEP

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EDIT LIMIT LINE

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EDIT LIST

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ELECTRICAL DELAY

SCALE REF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EMIT BEEP

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

END OF LABEL

DISPLA Y

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

END SWEEP HIGH PULSE

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

END SWEEP LOW PULSE

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ERASE TITLE

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ERASE TITLE

DISPLA Y

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ERASE TITLE

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EXT SOURCE AUTO

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EXT SOURCE MANUAL

SYSTEM

9-74 K ey De nitions

Show 30 more labels / short text on this page
Softkey Front-P anel · · · · · · · · · · · · · · · · · · · · · · · · · · · · ·
Page 561 of the Agilent Technologies 8753D user manual — SYSTEM
Page 561 · click to enlarge
Page 561

T able 9-2. Softkey Locations (continued)

A ccess K ey

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EXT TRIG ON POINT

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EXT TRIG ON SWEEP

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EXTENSION INPUT A

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EXTENSION INPUT B

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EXTENSION PORT 1

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EXTENSION PORT 2

CAL

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EXTENSIONS on OFF

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

EXTERNAL DISK

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FILETITLE FILE0

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNN

FILENAME

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FILE UTILITES

SA VE/RECALL

NNNNNNNNNNNNNNNNN

FIXED

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FIXED MKR AUX VALUE

MARKER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FIXED MKR POSITION

MARKER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FIXED MKR STIMULUS

MARKER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FIXED MKR VALUE

MARKER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FLAT LINE

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FORM FEED

DISPLA Y

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FORMAT ARY on OFF

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FORMAT DISK

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FORMAT: DOS

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FORMAT: LIF

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FORMAT EXT DISK

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FORMAT INT DISK

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FORMAT INT MEMORY

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FREQ OFFS MENU

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FREQ OFFS on OFF

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FREQUENCY

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FREQUENCY BLANK

DISPLA Y

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FREQUENCY: CW

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FREQUENCY: SWEEP

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FULL 2-PORT

CAL

K ey De nitions 9-75

Show 33 more labels / short text on this page
Softkey Front-P anel · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ·
Page 562 of the Agilent Technologies 8753D user manual — LOCAL
Page 562 · click to enlarge
Page 562

T able 9-2. Softkey Locations (continued)

A ccess K ey

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FULL PAGE

COPY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FWD ISOL'N ISOL'N STD

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FWD MATCH

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FWD MATCH THRU

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FWD TRANS

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

FWD TRANS THRU

CAL

N

NNNNNNNNNNNNNNNNNNNNNNNNN

MARKER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GATE: CENTER

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GATE: SPAN

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GATE: START

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GATE: STOP

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GATE on OFF

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GATE SHAPE

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GATE SHAPE MAXIMUM

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GATE SHAPE MINIMUM

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GATE SHAPE NORMAL

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GOSUB SEQUENCE

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GRAPHICS on OFF

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GRATICULE [ ]

COPY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GRATICULE TEXT

DISPLA Y

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

HARMONIC MEAS

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

HARMONIC OFF

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

HARMONIC SECOND

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

HARMONIC THIRD

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

HELP ADAPT REMOVAL

CAL

NNNNNNNNNNNNNN

HOLD

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

HP-IB DIAG on off

LOCAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

IF BW [ ]

A V G

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

IF LIMIT TEST FAIL

SEQ

N

IF LIMIT TEST PASS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

IF LOOP COUNTER = 0

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

IF LOOP < > COUNTER 0

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

IMAGINARY

F ORMA T

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INCR LOOP COUNTER

SEQ

9-76 K ey De nitions

Show 36 more labels / short text on this page
Softkey Front-P anel · · · · · · · G+jB MKR · · · · · · · · · · · · · · · · · · · · · · · · · · · ·
Page 563 of the Agilent Technologies 8753D user manual — SYSTEM
Page 563 · click to enlarge
Page 563

T able 9-2. Softkey Locations (continued)

A ccess K ey

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INIT DISK? YES

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INITIALIZE DISK

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INPUT PORTS

MEAS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INSTRUMENT MODE

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INTENSITY

DISPLA Y

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INTERNAL DISK

SA VE/RECALL

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INTERNAL MEMORY

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

INTERPOL on OFF

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ISOLATION

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ISOLATION DONE

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ISOL'N STD

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ISTATE CONTENTS

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

KIT DONE (MODIFIED)

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LABEL CLASS

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LABEL CLASS DONE

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LABEL KIT

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LABEL STD

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LEFT LOWER

COPY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LEFT UPPER

COPY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LIMIT LINE OFFSETS

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LIMIT LINE on OFF

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LIMIT MENU

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LIMIT TEST on OFF

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LIMIT TEST RESULT

DISPLA Y

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LIMIT TYPE

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNN

LIN FREQ

MENU

NNNNNNNNNNNNNNNNNNNNNNN

LIN MAG

F ORMA T

NNNNNNNNNNNNNNNNNNNNNNN

LIN MKR

MARKER F CTN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LIST FREQ

MENU

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LINE/MATCH

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LINE TYPE DATA

COPY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LINE TYPE MEMORY

COPY

NNNNNNNNNNNNNN

LIST

COPY

K ey De nitions 9-77

Show 34 more labels / short text on this page
Softkey Front-P anel · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ·
Page 564 of the Agilent Technologies 8753D user manual — SYSTEM
Page 564 · click to enlarge
Page 564

T able 9-2. Softkey Locations (continued)

A ccess K ey

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LN/MATCH 1

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LN/MATCH 2

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LO CONTROL on OFF

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNN

LO MENU

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LO SOURCE ADDRESS

SYSTEM

NNNNNNNNNNNNNN

LOAD

CAL

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOAD NO OFFSET

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOAD OFFSET

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOAD SEQ FROM DISK

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNN

LOG FREQ

MENU

NNNNNNNNNNNNNNNNNNNNNNN

LOG MAG

F ORMA T

NNNNNNNNNNNNNNNNNNNNNNN

LOG MKR

MARKER F CTN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOOP COUNTER

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOOP COUNTER

DISPLA Y

NNNNNNNNNNNNNN

LOSS

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOSS/SENSR LISTS

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOWER LIMIT

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOW PASS IMPULSE

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOW PASS STEP

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MANUAL TRG ON POINT

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER ! AMP. OFS.

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER ! CENTER

MARKER F CTN

9-78 K ey De nitions

Show 23 more labels / short text on this page
Softkey Front-P anel · · · · · · · · · · · · · · · · · · · · · ·
Page 565 of the Agilent Technologies 8753D user manual — MARKER
Page 565 · click to enlarge
Page 565

T able 9-2. Softkey Locations (continued)

A ccess K ey

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER ! CW

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER ! DELAY

MARKER F CTN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER ! DELAY

SCALE REF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER ! MIDDLE

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER ! REFERENCE

MARKER F CTN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER ! REFERENCE

SCALE REF

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER ! SPAN

MARKER F CTN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER ! START

MARKER F CTN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER ! STIMULUS

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER ! STOP

MARKER F CTN

NNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER 1

MARKER

NNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER 2

MARKER

NNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER 3

MARKER

NNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER 4

MARKER

NNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER 5

MARKER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER all OFF

MARKER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER MODE MENU

MARKER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKERS: CONTINUOUS

MARKER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKERS: COUPLED

MARKER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKERS: DISCRETE

MARKER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKERS: UNCOUPLED

MARKER

NNNNNNNNNNN

MAX

MARKER F CTN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MAXIMUM FREQUENCY

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MEASURE RESTART

MENU

NNNNNNNNNNNNNNNNNNNN

MEMORY

DISPLA Y

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MIDDLE VALUE

SYSTEM

NNNNNNNNNNN

MIN

MARKER F CTN

NNNNNNNNNNNNNNNNNNNNNNN

MINIMUM

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MINIMUM FREQUENCY

CAL

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MKR SEARCH [ ]

MARKER F CTN

NNNNNNNNNNNNNNNNNNNNNNNNNN

MKR ZERO

MARKER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MODIFY [ ]

CAL

K ey De nitions 9-79

Show 33 more labels / short text on this page
Softkey Front-P anel · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ·
Page 566 of the Agilent Technologies 8753D user manual — COPY
Page 566 · click to enlarge
Page 566

T able 9-2. Softkey Locations (continued)

A ccess K ey

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MODIFY COLORS

DISPLA Y

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NETWORK ANALYZER

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NEW SEQ/MODIFY SEQ

SEQ

NNNNNNNNNNNNNNNNNNNNNNN

NEWLINE

DISPLA Y

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NEXT PAGE

COPY

NNNNNNNNNNNNNNNNNNNN

NORMAL

SYSTEM

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NUMBER OF GROUPS

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NUMBER OF POINTS

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NUMBER OF READINGS

CAL

NNNNNNNNNNNNNNNNNNNN

OFFSET

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

OFFSET DELAY

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

OFFSET LOADS DONE

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

OFFSET LOSS

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

OFFSET Z0

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

OMIT ISOLATION

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ONE-PATH 2-PORT

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ONE SWEEP

CAL

NNNNNNNNNNNNNN

OPEN

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

OP PARMS (MKRS etc)

COPY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

P MTR/HPIB TO TITLE

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PARALL IN BIT NUMBER

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PARALL IN IF BIT H

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PARALL IN IF BIT L

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNN

PARALLEL

LOCAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PARALLEL [ ]

LOCAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PARALLEL OUT ALL

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PAUSE TO SELECT

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PEN NUM DATA

COPY

9-80 K ey De nitions

Show 29 more labels / short text on this page
Softkey Front-P anel · · · · · · · · · · · · · · · · · · · · · · · · · · · ·
Page 567 of the Agilent Technologies 8753D user manual — COPY
Page 567 · click to enlarge
Page 567

T able 9-2. Softkey Locations (continued)

A ccess K ey

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PEN NUM GRATICULE

COPY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PEN NUM MARKER

COPY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PEN NUM MEMORY

COPY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PEN NUM TEXT

COPY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PERIPHERAL HPIB ADDR

SEQ

NNNNNNNNNNNNNNNNN

PHASE

F ORMA T

N

NNNNNNNNNNNNNNNN

PHASE

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PHASE OFFSET

SCALE REF

NNNNNNNNNNNNNN

PLOT

COPY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOT DATA ON off

COPY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOT GRAT ON off

COPY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOT MEM ON off

COPY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOT MKR ON off

COPY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOT SPEED [ ]

COPY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOT TEXT ON off

COPY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOTTER BAUD RATE

LOCAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOTTER FORM FEED

COPY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOTTER PORT

LOCAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLTR PORT: DISK

LOCAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLTR PORT: HPIB

LOCAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLTR PORT: PARALLEL

LOCAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLTR PORT: SERIAL

LOCAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLTR TYPE [ ]

LOCAL

NNNNNNNNNNNNNNNNN

POLAR

F ORMA T

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

POLAR MKR MENU

MARKER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PORT EXTENSIONS

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PORT PWR [COUPLED]

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PORT PWR [UNCOUPLED]

MENU

NNNNNNNNNNNNNNNNN

POWER

MENU

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

POWER: FIXED

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

POWER: SWEEP

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

POWER LOSS

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

POWER MTR [ ]

LOCAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

POWER RANGES

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

POWER SWEEP

MENU

K ey De nitions 9-81

Show 36 more labels / short text on this page
Softkey Front-P anel · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ·
Page 568 of the Agilent Technologies 8753D user manual — CAL
Page 568 · click to enlarge
Page 568

T able 9-2. Softkey Locations (continued)

A ccess K ey

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRESET: FACTORY

PRESET

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRESET: USER

PRESET

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PREVIOUS PAGE

COPY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRINT: COLOR

COPY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRINT COLORS

COPY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRINT: MONOCHROME

COPY

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRINT MONOCHROME

COPY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRINT SEQUENCE

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRINTER BAUD RATE

LOCAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRINTER FORM FEED

COPY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRINTER PORT

LOCAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRNTR PORT: HPIB

LOCAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRNTR PORT: PARALLEL

LOCAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRNTR PORT: SERIAL

LOCAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRNTR TYPE [ ]

LOCAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PWR LOSS on OFF

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PWR RANGE AUTO man

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PWRMTR CAL [ ]

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PWRMTR CAL [OFF]

CAL

NNNNN

R

MEAS

NNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RANGE 0 -15 TO +10

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RANGE 1 -25 TO 0

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RANGE 2 -35 TO -10

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RANGE 3 -45 TO -20

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RANGE 4 -55 TO -30

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RANGE 5 -65 TO -40

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RANGE 6 -75 TO -50

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RANGE 7 -85 TO -60

MENU

9-82 K ey De nitions

Show 31 more labels / short text on this page
Softkey Front-P anel · · · · · · · · · · · · · · · · · · · · · R+jX MKR · · · · · · · · ·
Page 569 of the Agilent Technologies 8753D user manual — SA VE/RECALL
Page 569 · click to enlarge
Page 569

T able 9-2. Softkey Locations (continued)

A ccess K ey

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RAW ARRAY on OFF

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RAW OFFSET ON Off

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER

NNNNNNNNNNNNNN

REAL

F ORMA T

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECALL CAL PORT 1

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECALL CAL PORT 2

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECALL CAL SETS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECALL COLORS

DISPLA Y

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECALL KEYS MENU

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECALL KEYS on OFF

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECALL REG1

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECALL REG2

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECALL REG3

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECALL REG4

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECALL REG5

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECALL REG6

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECALL REG7

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECALL STATE

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RECEIVER CAL

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REFERENCE POSITION

SCALE REF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REFERENCE VALUE

SCALE REF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REFL: FWD S11 (A/R)

MEAS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REFL: REV S22 (B/R)

MEAS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REFLECT AND LINE

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REFLECTION

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REMOVE ADAPTER

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RENAME FILE

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RE-SAVE STATE

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RESET COLOR

DISPLA Y

N

NNNNNNNNNNNNNNNNNNNNNNNNN

RESPONSE

CAL

K ey De nitions 9-83

Show 35 more labels / short text on this page
Softkey Front-P anel · · · Re/Im MKR · · · Cal · · Cal · · Cal · · · · · · · · · · · · · · · · · · · · · · · ·
Page 570 of the Agilent Technologies 8753D user manual — SYSTEM
Page 570 · click to enlarge
Page 570

T able 9-2. Softkey Locations (continued)

A ccess K ey

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RESPONSE & ISOL'N

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RESUME CAL SEQUENCE

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REV ISOL'N ISOL'N STD

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REV MATCH

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REV MATCH THRU

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REV TRANS

CAL

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REV TRANS THRU

CAL

NNNNNNNNNNNNNNNNNNNNNNN

RF > LO

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNN

RF < LO

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RIGHT LOWER

COPY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RIGHT UPPER

COPY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ROUND SECONDS

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

S PARAMETERS

MEAS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

S11 1-PORT

CAL

NNNNNNNNNNNNNN

S11A

CAL

NNNNNNNNNNNNNN

S11B

CAL

NNNNNNNNNNNNNN

S11C

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

S11 REFL OPEN

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

S22 1-PORT

CAL

NNNNNNNNNNNNNN

S22A

CAL

NNNNNNNNNNNNNN

S22B

CAL

NNNNNNNNNNNNNN

S22C

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

S22 REFL OPEN

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SAMPLR COR ON off

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SAVE COLORS

DISPLA Y

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SAVE USER KIT

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SAVE USING ASCII

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SAVE USING BINARY

SA VE/RECALL

9-84 K ey De nitions

Show 29 more labels / short text on this page
Softkey Front-P anel · · · · · · · · · · · · · · · · · · · · · · · · · · · ·
Page 571 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 571 · click to enlarge
Page 571

T able 9-2. Softkey Locations (continued)

A ccess K ey

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SCALE/DIV

SCALE REF

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SCALE PLOT [ ]

COPY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEARCH LEFT

MARKER F CTN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEARCH RIGHT

MARKER F CTN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEARCH: MAX

MARKER F CTN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEARCH: MIN

MARKER F CTN

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEARCH: OFF

MARKER F CTN

NNNNNNNNNNNNNNNNNNNN

SECOND

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNN

SEGMENT

CAL

NNNNNNNNNNNNNNNNNNNNNNN

SEGMENT

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEGMENT: CENTER

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEGMENT: SPAN

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEGMENT: START

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEGMENT: STOP

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEL QUAD [ ]

COPY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SELECT DISK

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQUENCE 1 SEQ1

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQUENCE 2 SEQ2

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQUENCE 3 SEQ3

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQUENCE 4 SEQ4

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQUENCE 5 SEQ5

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQUENCE 6 SEQ6

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SEQUENCE FILENAMING

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SET ADDRESSES

LOCAL

NNNNNNNNNNNNNNNNNNNNNNN

SET BIT

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SET CLOCK

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNN

SET DAY

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SET FREQ LOW PASS

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNN

SET HOUR

SYSTEM

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SET MINUTES

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SET MONTH

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SET REF: THRU

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SET REF: REFLECT

SYSTEM

K ey De nitions 9-85

Show 35 more labels / short text on this page
Softkey Front-P anel · · · · · · · · · · · · · · · · · · · · · · · Save/Recall · · · · · · · · · · ·
Page 572 of the Agilent Technologies 8753D user manual — MARKER F CTN
Page 572 · click to enlarge
Page 572

T able 9-2. Softkey Locations (continued)

A ccess K ey

NNNNNNNNNNNNNNNNNNNNNNNNNN

SET YEAR

SYSTEM

NNNNNNNNNNNNNNNNNNNN

SET Z0

CAL

NNNNNNNNNNNNNNNNN

SHORT

CAL

NNNNNNNNNNNNNNNNNNNN

SINGLE

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SINGLE POINT

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SINGLE SEG SWEEP

MENU

N

NNNNNNNNNNNNNNNNNNNNNN

SLIDING

CAL

NNNNNNNNNNNNNNNNN

SLOPE

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SLOPE on OFF

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SLOPING LINE

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SMITH CHART

F ORMA T

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SMITH MKR MENU

MARKER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SMOOTHING APERTURE

A V G

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SMOOTHING on OFF

A V G

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SOURCE PWR ON off

MENU

NNNNNNNNNNNNNN

SPAN

MENU

NNNNNNNNNNNNNN

SPAN

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SPECIAL FUNCTIONS

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SPECIFY CLASS

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SPECIFY GATE

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SPECIFY OFFSET

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SPLIT DISP on OFF

DISPLA Y

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SPUR AVOID On Off

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STANDARDS DONE

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STATS on OFF

MARKER F CTN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STD DONE (MODIFIED)

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STD OFFSET DONE

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STD TYPE:

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STEP SIZE

MENU

N

STIMULUS VALUE

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STIMULUS OFFSET

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

STORE SEQ TO DISK

SEQ

9-86 K ey De nitions

Show 33 more labels / short text on this page
Softkey Front-P anel · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ·
Page 573 of the Agilent Technologies 8753D user manual — SYSTEM
Page 573 · click to enlarge
Page 573

T able 9-2. Softkey Locations (continued)

A ccess K ey

NNNNNNNNNNNNNNNNN

SWEEP

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SWEEP TIME [ ]

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SWEEP TYPE MENU

MENU

NNNNNNNNNNN

SWR

F ORMA T

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SYSTEM CONTROLLER

LOCAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TAKE CAL SWEEP

CAL

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TAKE RCVR CAL SWEEP

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TALKER/LISTENER

LOCAL

NNNNNNNNNNNNNNNNNNNN

TARGET

MARKER F CTN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TERMINAL IMPEDANCE

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TEST PORT 1 2

MEAS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TESTSET I/O FWD

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TESTSET I/O REV

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TESTSET SW XXXX

SYSTEM

NNNNNNNNNNNNNN

TEXT

DISPLA Y

NNNNNNNNNNNNNNNNNNNNNNNNNN

TEXT [ ]

COPY

NNNNNNNNNNNNNN

THRU

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

THRU THRU

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TIME STAMP ON off

SYSTEM

NNNNNNNNNNNNNN

TINT

DISPLA Y

NNNNNNNNNNNNNNNNN

TITLE

DISPLA Y

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TITLE SEQUENCE

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TITLE TO MEMORY

SEQ

K ey De nitions 9-87

Show 26 more labels / short text on this page
Softkey Front-P anel · · · · · · · · · · · · · · · · Cal · · · · · · · · ·
Page 574 of the Agilent Technologies 8753D user manual — SYSTEM
Page 574 · click to enlarge
Page 574

T able 9-2. Softkey Locations (continued)

A ccess K ey

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TITLE TO P MTR/HPIB

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TITLE TO PERIPHERAL

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TITLE TO PRNTR/HPIB

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRACKING on OFF

MARKER F CTN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRANS DONE

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRANS: FWD S21 (B/R)

MEAS

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRANS: REV S12 (B/R)

MEAS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRANSFORM MENU

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRANSFORM on OFF

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRANSMISSION

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRIGGER MENU

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRIGGER: TRIG OFF

MENU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRL*/LRM* 2-PORT

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TRL/LRM OPTION

CAL

NNNNNNNNNNNNNNNNNNNNNNN

TTL I/O

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TTL OUT HIGH

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TTL OUT LOW

SEQ

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TUNED RECEIVER

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

UNCOUPLED

MARKER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

UP CONVERTER

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

UPPER LIMIT

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

USE MEMORY on OFF

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

USE PASS CONTROL

LOCAL

NNNNNNNNNNNNNN

USER

PRESET

NNNNNNNNNNNNNNNNNNNNNNNNNN

USER KIT

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

USE SENSOR A / B

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

VELOCITY FACTOR

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

VIEW MEASURE

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

VOLUME NUMBER

LOCAL

N

VOLUME NUMBER

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SA VE/RECALL

NNNNNNNNNNNNNNNNNNNN

SEQ

NNNNNNNNNNNNNNNNNNNNNNN

WARNING

DISPLA Y

9-88 K ey De nitions

Show 34 more labels / short text on this page
Softkey Front-P anel · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · WAIT x · ·
Page 575 of the Agilent Technologies 8753D user manual — MARKER F CTN
Page 575 · click to enlarge
Page 575

T able 9-2. Softkey Locations (continued)

A ccess K ey

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

WARNING [ ]

COPY

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

WAVEGUIDE

CAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

WAVEGUIDE DELAY

SCALE REF

NNNNNNNNNNNNNN

WIDE

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

WIDTH VALUE

MARKER F CTN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

WIDTHS on OFF

MARKER F CTN

N

NNNNNNNNNNNNNNNNNNN

WINDOW

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

WINDOW: MAXIMUM

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

WINDOW: MINIMUM

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

WINDOW: NORMAL

SYSTEM

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

XMIT CNTRL [ ]

LOCAL

NNNNNNNNNNNNNNNNNNNNNNN

Y: REFL

MEAS

NNNNNNNNNNNNNNNNNNNNNNNNNN

Y: TRANS

MEAS

NNNNNNNNNNNNNNNNNNNNNNN

Z: REFL

MEAS

NNNNNNNNNNNNNNNNNNNNNNNNNN

Z: TRANS

MEAS

K ey De nitions 9-89

Show 16 more labels / short text on this page
Softkey Front-P anel · · · · · · · · · · · · · · ·
Page 576 of the Agilent Technologies 8753D user manual — Chapter 9 Key Definitions
Page 576 · click to enlarge
Page 576
Page 576 of the Agilent Technologies 8753D user manual — Chapter 9 Key Definitions
Page 577 of the Agilent Technologies 8753D user manual — Chapter 10 Error Messages
Page 577 · click to enlarge
Page 577

This chapter contains the following information to help you interpret any error messages that

may be displayed on the analyzer LCD or transmitted by the instrument over HP-IB:

An alphabetical listing of all error messages , including:

An explanation of the message

Suggestions to help solve the problem

A numerical listing of all error messages

Some messages described in this chapter are for information only and do not

indicate an error condition. These messages are not numbered and so they will

not appear in the numerical listing.

Where to Look for More Information

A dditional information about many of the topics discussed in this chapter is located in the

following areas:

Chapter 2, \Making Measurements ," contains step-by-step procedures for making

measurements or using particular functions .

Chapter 4, \Printing, Plotting, and Saving Measurement Results ," contains instructions

for saving to disk or the analyzer internal memory , and printing and plotting displayed

measurements .

Chapter 6, \Application and Operation Concepts ," contains explanatory-style information

about many applications and analyzer operation.

Chapter 9, \Key De nitions ," describes all the front panel keys , softkeys , and their

corresponding HP-IB commands .

Chapter 12, \Preset State and Memory Allocation," contains a discussion of memory

allocation, memory storage , instrument state de nitions , and preset conditions .

Error Messages · Error Messages 10-1

Page 578 of the Agilent Technologies 8753D user manual — ANOTHER SYSTEM CONTROLLER ON HP-IB BUS
Page 578 · click to enlarge
Page 578

Error Messages in Alphabetical Order

ABORTING COPY OUTPUT

This message is displayed brie y if you have pressed

to abort a copy

LOCAL

operation. If the message is not subsequently replaced by error message

number 25, PRINT ABORTED , the copy device may be hung. Press

LOCAL

more to exit the abort process and verify the status of the copy device . At this

point, the copy device will probably have an error condition which must be

xed. (F or exampleout of paper or paper jam.)

ADDITIONAL STANDARDS NEEDED

Error correction for the selected calibration class cannot be computed until you

have measured all the necessary standards .

ADDRESSED TO TALK WITH NOTHING TO SAY

Y ou have sent a read command to the analyzer (such as ENTER 716) without

rst requesting data with an appropriate output command (such as OUTPD A T A).

The analyzer has no data in the output queue to satisfy the request.

AIR FLOW RESTRICTED: CHECK FAN FILTER

Something is restricting the air ow into the analyzer . Check for any debris and

clean or replace the fan lter .

ALL REGISTERS HAVE BEEN USED

Y ou have used all of the available registers; you can store no more instrument

states even though you may still have sucient memory . There are 31 registers

available , plus the present instrument state .

ANOTHER SYSTEM CONTROLLER ON HP-IB BUS

Y ou must remove the active controller from the bus or the controller must

relinquish the bus before the analyzer can assume the system controller mode .

Show 9 more labels / short text on this page
Information · Message · once · Error Number · Error Number · Error Number · Error Number · Error Number · 10-2 Error Messages
Page 579 of the Agilent Technologies 8753D user manual — BATTERY FAILED. STATE MEMORY CLEARED
Page 579 · click to enlarge
Page 579

ASCII: MISSING 'BEGIN' STATEMENT

The citi le you just downloaded over the HP-IB or via disk was not properly

organized. The analyzer is unable to read the \BEGIN" statement.

ASCII: MISSING 'CITIFILE' STATEMENT

The citi le you just downloaded over the HP-IB or via disk was not properly

organized. The analyzer is unable to read the \CITIFILE" statement.

ASCII: MISSING 'DATA' STATEMENT

The citi le you just downloaded over the HP-IB or via disk was not properly

organized. The analyzer is unable to read the \D A T A" statement.

ASCII: MISSING 'VAR' STATEMENT

The citi le you just downloaded over the HP-IB or via disk was not properly

organized. The analyzer is unable to read the \V AR" statement.

AVERAGING INVALID ON NON-RATIO MEASURE

Y ou cannot use sweep-to-sweep averaging in single-input measurements .

Sweep-sweep averaging is valid only for ratioed measurements (A/R, B/R, A/B ,

and S-parameters). Y ou can use noise reduction techniques , such as narrower

IF bandwidth, for single input measurements .

BAD FREQ FOR HARMONIC OR FREQ OFFSET

Y ou turned on time domain or recalled a calibration that resulted in start and

stop frequencies that are beyond the allowable limits .

BATTERY FAILED. STATE MEMORY CLEARED

The battery protection of the non-volatile CMOS memory has failed. The CMOS

memory has been cleared. Refer to the HP 8753D Network Analyzer Service

Guide for battery replacement instructions . See Chapter 12, \Preset State and

Memory Allocation," for more information about the CMOS memory .

Show 8 more labels / short text on this page
Error Number · Error Number · Error Number · Error Number · Error Number · Error Number · Error Number · Error Messages 10-3
Page 580 of the Agilent Technologies 8753D user manual — CANNOT FORMAT DOS DISKS ON THIS DRIVE
Page 580 · click to enlarge
Page 580

BATTERY LOW! STORE SAVE REGS TO DISK

The battery protection of the non-volatile CMOS memory is in danger of failing.

If this occurs , all of the instrument state registers stored in CMOS memory will

be lost. Save these states to a disk and refer to the HP 8753D Network Analyzer

Service Guide for battery replacement instructions . See Chapter 12, \Preset

State and Memory Allocation," for more information about the CMOS memory .

BLOCK INPUT ERROR

The analyzer did not receive a complete data transmission. This is usually

caused by an interruption of the bus transaction. Clear by pressing the

LOCAL

key or aborting the I/O process at the controller .

BLOCK INPUT LENGTH ERROR

The length of the header received by the analyzer did not agree with the size

of the internal array block. Refer to the HP 8753D Network Analyzer

Programmer's Guide for instructions on using analyzer input commands .

CALIBRATION ABORTED

Y ou have changed the active channel during a calibration so the calibration in

progress was terminated. Make sure the appropriate channel is active and

restart the calibration.

CALIBRATION REQUIRED

A calibration set could not be found that matched the current stimulus state or

measurement parameter . Y ou will have to perform a new calibration.

CANNOT FORMAT DOS DISKS ON THIS DRIVE

Y ou have attempted to initialize a oppy disk to DOS format on an external

disk drive that does not support writing to all 80 tracks of the double density

and high density disks . The older single-sided disks had only 66 tracks and

some disk drives were limited to accessing that number of tracks . T o format

the disk, either choose another external disk drive or use the analyzer's

internal disk drive .

Show 8 more labels / short text on this page
Error Number · Error Number · · Error Number · Error Number · Error Number · Error Number · 10-4 Error Messages
Page 581 of the Agilent Technologies 8753D user manual — CONTINUOUS SWITCHING NOT ALLOWED
Page 581 · click to enlarge
Page 581

CANNOT MODIFY FACTORY PRESET

Y ou have attempted to rename , delete , or otherwise alter the factory preset

state . The factory preset state is permanently stored in CMOS memory and

cannot be altered. If your intent was to create a user preset state , you must

create a new instrument state , save it, and then rename it to \UPRESET".

Refer to Chapter 12, \Preset State and Memory Allocation," for more detailed

instructions .

CANNOT READ/WRITE HFS FILE SYSTEM

The disk is being accessed by the analyzer and is found to contain an HFS

(hierarchical le system) or les nested within subdirectories . The analyzer

does not support HFS. Replace the disk medium with a LIF or DOS formatted

disk that does not contain les nested within subdirectories .

CAN'T STORE/LOAD SEQUENCE, INSUFFICIENT MEMORY

Y our sequence transfer to or from a disk could not be completed due to

insucient memory .

CH1 (CH2) TARGET VALUE NOT FOUND

Y our target value for the marker search function does not exist on the current

data trace .

CONTINUOUS SWITCHING NOT ALLOWED

Y our current measurement requires di erent power ranges on channel 1 and

channel 2. T o protect the attenuator from undue mechanical wear , test set hold

will be activated.

The \tsH" (test set hold) indicator in the left margin of the display indicates

that the inactive channel has been put in the sweep hold mode .

COPYdevice not responding; copy aborted

The printer or plotter is not accepting data. V erify the cable connections , HP-IB

addresses , and otherwise ensure that the copy device is ready .

Show 7 more labels / short text on this page
Error Number · Error Number · Error Number · Error Number · Error Number · Error Number · Error Messages 10-5
Page 582 of the Agilent Technologies 8753D user manual — D2/D1 INVALID WITH SINGLE CHANNEL
Page 582 · click to enlarge
Page 582

COPY OUTPUT COMPLETED

The analyzer has completed outputting data to the printer or plotter . The

analyzer can now accept another copy command.

CORRECTION AND DOMAIN RESET

When you change the frequency range , sweep type , or number of points ,

error-correction is switched o and the time domain transform is recalculated,

without error-correction. Y ou can either correct the frequency range , sweep

type , or number of points to match the calibration, or perform a new

calibration. Then perform a new time domain transform.

CORRECTION CONSTANTS NOT STORED

A store operation to the EEPROM was not successful. Y ou must change the

position of the jumper on the A9 CPU assembly . Refer to the \A9 CC Jumper

P osition Procedure" in the \A djustments and Correction Constants" chapter of

the HP 8753D Network Analyzer Service Guide.

CORRECTION TURNED OFF

Critical parameters in your current instrument state do not match the

parameters for the calibration set, therefore correction has been turned o .

The critical instrument state parameters are sweep type , start frequency ,

frequency span, and number of points .

CURRENT PARAMETER NOT IN CAL SET

Correction is not valid for your selected measurement parameter . Either

change the measurement parameters or perform a new calibration.

D2/D1 INVALID WITH SINGLE CHANNEL

Y ou can only make a D2/D1 measurement if both channels are on.

Show 8 more labels / short text on this page
Information · Message · Error Number · Error Number · Error Number · Error Number · Error Number · 10-6 Error Messages
Page 583 of the Agilent Technologies 8753D user manual — DISK HARDWARE PROBLEM
Page 583 · click to enlarge
Page 583

D2/D1 INVALID: CH1 CH2 NUM PTS DIFFERENT

Y ou can only make a D2/D1 measurement if both channels have the same

number of points .

DEADLOCK

A fatal rmware error occurred before instrument preset completed. Call your

local Hewlett-P ackard sales and service oce .

DEMODULATION NOT VALID

Demodulation was selected when the analyzer was not in CW time mode .

Select demodulation only after putting the analyzer into CW time mode .

DEVICEnot on, not connect, wrong addrs

The device at the selected address cannot be accessed by the analyzer . V erify

that the device is switched on, and check the HP-IB connection between the

analyzer and the device . Ensure that the device address recognized by the

analyzer matches the HP-IB address set on the device itself.

DIRECTORY FULL

There is no room left in the directory to add les . Either delete les or get a

new disk.

DISK HARDWARE PROBLEM

The disk drive is not responding correctly . Refer to the HP 8753D Network

Analyzer Service Guide for troubleshooting information. If using an external

disk drive , refer to the disk drive operating manual.

Show 7 more labels / short text on this page
Error Number · Error Number · Error Number · Error Number · Error Number · Error Number · Error Messages 10-7
Page 584 of the Agilent Technologies 8753D user manual — DOMAIN RESET
Page 584 · click to enlarge
Page 584

DISK IS WRITE PROTECTED

The store operation cannot write to a write-protected disk. Slide the

write-protect tab over the write-protect opening in order to write data on the

disk.

DISK MEDIUM NOT INITIALIZED

Y ou must initialize the disk before it can be used.

DISK MESSAGE LENGTH ERROR

The analyzer and the external disk drive aren't communicating properly . Check

the HP-IB connection and then try substituting another disk drive to isolate the

problem instrument.

DISKnot on, not connected, wrong addrs

The disk cannot be accessed by the analyzer . V erify power to the disk drive ,

and check the HP-IB connection between the analyzer and the disk drive .

Ensure that the disk drive address recognized by the analyzer matches the

HP-IB address set on the disk drive itself.

DISK READ/WRITE ERROR

There may be a problem with your disk. Try a new oppy disk. If a new oppy

disk does not eliminate the error , suspect hardware problems .

DISK WEAR - REPLACE DISK SOON

Cumulative use of the disk is approaching the maximum. Copy les as

necessary using an external controller . If no controller is available , load

instrument states from the old disk and store them to a newly initialized disk

using the save/recall features of the analyzer . Discard the old disk.

DOMAIN RESET

Time domain calculations were reset due to a change in the frequency range ,

sweep type , or number of points . P erform a new time domain transform on the

new state .

Show 8 more labels / short text on this page
Error Number · Error Number · Error Number · Error Number · Error Number · Error Number · Error Number · 10-8 Error Messages
Page 585 of the Agilent Technologies 8753D user manual — EXT SRC: NOT ON/CONNECTED OR WRONG ADDR
Page 585 · click to enlarge
Page 585

DOS NAME LIMITED TO 8 CHARS + 3 CHAR EXTENSION

A DOS le name must meet the following criteria:

minimum of 1 character

format is filename.ext

maximum of 8 characters in the lename

maximum of 3 characters in the extension eld (optional)

a dot separates the lename from the extension eld (the dot is not part of

the name on the disk)

DUPLICATING TO THIS SEQUENCE NOT ALLOWED

A sequence cannot be duplicated to itself.

EXCEEDED 7 STANDARDS PER CLASS

When modifying calibration kits , you can de ne a maximum of seven standards

for any class .

EXTERNAL SOURCE MODE REQUIRES CW TIME

An external source can only be phase locked and measured in the CW time

sweep mode .

EXT SOURCE NOT READY FOR TRIGGER

There is a hardware problem with the HP 8625A external source . V erify the

connections between the analyzer and the external source . If the connections

are correct, refer to the source operating manual.

EXT SRC: NOT ON/CONNECTED OR WRONG ADDR

The analyzer is unable to communicate with the external source . Check the

connections and the HP-IB address on the source .

Show 7 more labels / short text on this page
Error Number · Error Number · Error Number · Error Number · Error Number · Error Number · Error Messages 10-9
Page 586 of the Agilent Technologies 8753D user manual — FREQ OFFSET ONLY VALID IN NETWORK ANALYZER MODE
Page 586 · click to enlarge
Page 586

FILE NOT COMPATIBLE WITH INSTRUMENT

Y ou cannot recall user graphics that had been saved on an earlier model of

analyzer with a monochrome display . These les cannot be used with the

HP 8753D .

FILE NOT FOUND

The requested le was not found on the current disk medium.

FILE NOT FOUND OR WRONG TYPE

During a resave operation, either the le was not found or the type of le was

not an instrument state le .

FIRST CHARACTER MUST BE A LETTER

The rst character of a disk le title or an internal save register title must be

an alpha character .

FORMAT NOT VALID FOR MEASUREMENT

Conversion measurements (Z or Y re ection and transmission) are not valid

with Smith chart and SWR formats .

FORMATTING DATA

The list information is being processed for a list data output to a copy device

and stored in the copy spool bu er . During this time , the analyzer's resources

are dedicated to this task (which takes less than a few seconds).

FREQ OFFSET ONLY VALID IN NETWORK ANALYZER MODE

Y ou can only make frequency o set measurements in the network analyzer

mode .

Show 10 more labels / short text on this page
Information · Message · Error Number · Error Number · Error Number · Error Number · Information · Message · Error Number · 10-10 Error Messages
Page 587 of the Agilent Technologies 8753D user manual — HP 8753 SOURCE PARAMETERS CHANGED
Page 587 · click to enlarge
Page 587

FREQS CANNOT BE CHANGED, TOO MANY POINTS

The number of points selected for setting the low pass transform frequencies is

too high. Reduce the number of points so that the low pass criteria is met.

FUNCTION NOT AVAILABLE

The function you requested over HP-IB is not available on the current

instrument.

FUNCTION NOT VALID

The function you requested is incompatible with the current instrument state .

FUNCTION NOT VALID DURING MOD SEQUENCE

Y ou cannot perform sequencing operations while a sequence is being modi ed.

FUNCTION NOT VALID FOR INTERNAL MEMORY

The function you selected only works with disk les .

FUNCTION ONLY VALID DURING MOD SEQUENCE

Y ou can only use the

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

GOSUB SEQUENCE capability when you are building a

sequence . Attempting to use this softkey at any other time returns an error

message and no action is taken.

HP 8753 SOURCE PARAMETERS CHANGED

Some of the stimulus parameters of the instrument state have been changed,

because you have turned correction on. A calibration set for the current

measurement parameter was found and activated. The instrument state was

updated to match the stimulus parameters of the calibration state .

This message also appears when you have turned on harmonic mode or

frequency o set and the present frequency range cannot be used with one of

these modes .

Show 8 more labels / short text on this page
Error Number · Error Number · Error Number · Error Number · Error Number · Error Number · Error Number · Error Messages 10-11
Page 588 of the Agilent Technologies 8753D user manual — INSUFFICIENT MEMORY FOR PRINT/PLOT
Page 588 · click to enlarge
Page 588

HPIB COPY IN PROGRESS, ABORT WITH LOCAL

An HP-IB copy was already in progress when you requested the HP-IB for

another function. T o abort the rst copy , press

, otherwise the HP-IB is

LOCAL

unavailable until the rst copy is completed.

ILLEGAL UNIT OR VOLUME NUMBER

The disk unit or volume number set in the analyzer is not valid. Refer to the

disk drive operating manual.

INIT DISK removes all data from disk

Continuing with the initialize operation will destroy any data currently on the

disk.

INITIALIZATION FAILED

The disk initialization failed, probably because the disk is damaged.

INSTRUMENT STATE MEMORY CLEARED

All instrument state registers have been cleared from memory along with any

saved calibration data, memory traces , and calibration kit de nitions .

A dditionally , all user-settable selections (such as HP-IB addresses) are set to

their defaults .

INSUFFICIENT MEMORY

Y our last front panel or HP-IB request could not be implemented due to

insucient memory space . In some cases , this is a fatal error from which you

can escape only by presetting the instrument.

INSUFFICIENT MEMORY FOR PRINT/PLOT

There is not enough memory available for the print or plot function. Increase

the available memory by changing or eliminating a memory-intensive operation

such as reducing the number of points in the sweep .

Show 9 more labels / short text on this page
Error Number · Error Number · Information · Message · Error Number · Error Number · Error Number · Error Number · 10-12 Error Messages
Page 589 of the Agilent Technologies 8753D user manual — MEMORY FOR CURRENT SEQUENCE IS FULL
Page 589 · click to enlarge
Page 589

INSUFFICIENT MEMORY, PWR MTR CAL OFF

There is not enough memory space for the power meter calibration array .

Increase the available memory by clearing one or more save/recall registers , or

by reducing the number of points .

INVALID KEY

Y ou pressed an unde ned softkey .

LIST MODE OFF: INVALID WITH LO FREQ

List mode has been turned o in the frequency o set mode because it is

incompatible with your selected LO frequency .

LIST TABLE EMPTY

The frequency list is empty . T o implement list frequency mode , add segments

to the list table .

LOG SWEEP REQUIRES 2 OCTAVE MINIMUM SPAN

A logarithmic sweep is only valid if the stop frequency is greater than four

times the start frequency . F or frequency spans of less than two octaves , the

sweep type automatically reverts to linear sweep .

LOW PASS: FREQ LIMITS CHANGED

The frequency domain data points must be harmonically related from dc to the

stop frequency . That is , stop = n start, where n = number of points . If this

condition is not true when a low pass mode (step or impulse) is selected and

transform is turned on, the analyzer resets the start and stop frequencies . The

stop frequency is set close to the entered stop frequency , and the start

frequency is set equal to stop/n.

MEMORY FOR CURRENT SEQUENCE IS FULL

All the memory in the sequence you are modifying is lled with instrument

commands .

Show 9 more labels / short text on this page
Error Number · Error Number · Error Number · Error Number · Error Number · Information · Message · Error Number · Error Messages 10-13
Page 590 of the Agilent Technologies 8753D user manual — NO LIMIT LINES DISPLAYED
Page 590 · click to enlarge
Page 590

MORE SLIDES NEEDED

When you use a sliding load (in a user-de ned calibration kit), you must set at

least three slide positions to complete the calibration.

NO CALIBRATION CURRENTLY IN PROGRESS

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

RESUME CAL SEQUENCE softkey is not valid unless a calibration is already

in progress . Start a new calibration.

NO DISK MEDIUM IN DRIVE

Y ou have no disk in the current disk unit. Insert a disk, or check the disk unit

number stored in the analyzer .

NO FAIL FOUND

The self-diagnose function of the instrument operates on an internal test

failure . At this time , no failure has been detected.

NO FILE(S) FOUND ON DISK

No les of the type created by an analyzer store operation were found on the

disk or the disk drive is empty . If you requested a speci c le title , that le

was not found on the disk.

NO IF FOUND: CHECK R INPUT LEVEL

The rst IF signal was not detected during pretune . Check the front panel R

channel jumper . If there is no visible problem with the jumper , refer to the

HP 8753D Network Analyzer Service Guide for troubleshooting.

NO LIMIT LINES DISPLAYED

Y ou can turn limit lines on but they cannot be displayed on polar or Smith

chart display formats .

Show 10 more labels / short text on this page
Error Number · Error Number · The · Error Number · Service Error · Number 114 · Error Number · Error Number · Error Number · 10-14 Error Messages
Page 591 of the Agilent Technologies 8753D user manual — NOT ENOUGH SPACE ON DISK FOR STORE
Page 591 · click to enlarge
Page 591

NO MARKER DELTA -SPAN NOT SET

Y ou must turn the delta marker mode on, with at least two markers displayed,

in order to use the

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

MARKER -> SPAN softkey function.

NO MEMORY AVAILABLE FOR INTERPOLATION

Y ou cannot perform interpolated error correction due to insucient memory .

NO MEMORY AVAILABLE FOR SEQUENCING

Y ou cannot modify the sequence due to insucient memory .

NO PHASE LOCK: CHECK R INPUT LEVEL

The rst IF signal was detected at pretune , but phase lock could not be

acquired. Check the signal level to the R channel input to make sure it is 35

dBm or higher . Refer to the HP 8753D Network Analyzer Service Guide for

troubleshooting.

NO SPACE FOR NEW CAL. CLEAR REGISTERS

Y ou cannot store a calibration set due to insucient memory . Y ou can free

more memory by clearing a saved instrument state from an internal register

(which may also delete an associated calibration set, if all the instrument states

using the calibration kit have been deleted). Y ou can store the saved

instrument state and calibration set to a disk before clearing them. After

deleting the instrument states , press

PRESET

to run the memory packer .

NOT ALLOWED DURING POWER METER CAL

When the analyzer is performing a power meter calibration, the HP-IB bus is

unavailable for other functions such as printing or plotting.

NOT ENOUGH SPACE ON DISK FOR STORE

The store operation will over ow the available disk space . Insert a new disk or

purge les to create free disk space .

Show 8 more labels / short text on this page
Error Number · Error Number · Error Number · Error Number · Error Number · Error Number · Error Number · Error Messages 10-15
Page 592 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 592 · click to enlarge
Page 592

NO VALID MEMORY TRACE

If you are going to display or otherwise use a memory trace , you must rst

store a data trace to memory .

NO VALID STATE IN REGISTER

Y ou have requested the analyzer , over HP-IB (or by sequencing), to load an

instrument state from an empty internal register .

ONLY LETTERS AND NUMBERS ARE ALLOWED

Y ou can only use alpha-numeric characters (and underscores) in disk le titles

or internal save register titles . Other symbols are not allowed, except for the

\underscore" symbol.

OPTIONAL FUNCTION; NOT INSTALLED

The function you requested requires a capability provided by an option to the

standard analyzer . That option is not currently installed. (Refer to Chapter 1

for a description of the available options .)

OVERLOAD ON INPUT A, POWER REDUCED

See error number 57.

OVERLOAD ON INPUT B, POWER REDUCED

See error number 57.

OVERLOAD ON INPUT R, POWER REDUCED

Y ou have exceeded approximately +14 dBm at one of the test ports . The RF

output power is automatically reduced to 85 dBm. The annotation P+ appears

in the left margin of the display to indicate that the power trip function has

been activated. When this occurs , reset the power to a lower level, then toggle

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SOURCE PWR on OFF softkey to switch on the power again.

Show 9 more labels / short text on this page
Error Number · Error Number · Error Number · Error Number · Error Number · Error Number · Error Number · the · 10-16 Error Messages
Page 593 of the Agilent Technologies 8753D user manual — LOCAL
Page 593 · click to enlarge
Page 593

PARALLEL PORT NOT AVAILABLE FOR GPIO

Y ou have de ned the parallel port as COPY for sequencing in the HP-IB menu.

T o access the parallel port for general purpose I/O (GPIO), set the selection to

[GPIO].

PARALLEL PORT NOT AVAILABLE FOR COPY

Y ou have de ned the parallel port as general purpose I/O (GPIO) for

sequencing. The de nition was made under the

LOCAL

key menus . T o access

the parallel port for copy , set the selection to

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PARALLEL [COPY] .

PHASE LOCK CAL FAILED

An internal phase lock calibration routine is automatically executed at

power-on, preset, and any time a loss of phase lock is detected. This message

indicates that phase lock calibration was initiated and the rst IF detected, but

a problem prevented the calibration from completing successfully . Refer to the

HP 8753D Network Analyzer Service Guide and execute pretune correction test

This message may appear if you connect a mixer between the RF output and R

input before turning on frequency o set mode . Ignore it: it will go away when

you turn on frequency o set. This message may also appear if you turn on

frequency o set mode before you de ne the o set.

PHASE LOCK LOST

Phase lock was acquired but then lost. Refer to the HP 8753D Network

Analyzer Service Guide for troubleshooting information.

PLOT ABORTED

When you press the

key , the analyzer aborts the plot in progress .

LOCAL

PLOTTERnot on, not connect, wrong addrs

The plotter does not respond to control. V erify power to the plotter , and check

the HP-IB connection between the analyzer and the plotter . Ensure that the

plotter address recognized by the analyzer matches the HP-IB address set on

the plotter itself.

Show 7 more labels / short text on this page
Error Number · Error Number · Error Number · Error Number · Error Number · Error Number · Error Messages 10-17
Page 594 of the Agilent Technologies 8753D user manual — POWER UNLEVELED
Page 594 · click to enlarge
Page 594

PLOTTER NOT READY-PINCH WHEELS UP

The plotter pinch wheels clamp the paper in place . If you raise the pinch

wheels , the plotter indicates a \not ready" status on the bus .

POSSIBLE FALSE LOCK

Phase lock has been achieved, but the source may be phase locked to the

wrong harmonic of the synthesizer . P erform the source pretune correction

routine documented in the \A djustments and Correction Constants" chapter in

the HP 8753D Network Analyzer Service Guide.

POWER METER INVALID

The power meter indicates an out-of-range condition. Check the test setup .

POWER METER NOT SETTLED

Sequential power meter readings are not consistent. V erify that the equipment

is set up correctly . If so , preset the instrument and restart the operation.

POWER SUPPLY HOT!

The temperature sensors on the A8 post-regulator assembly have detected an

over-temperature condition. The power supplies regulated on the

post-regulator have been shut down.

POWER SUPPLY SHUT DOWN!

One or more supplies on the A8 post-regulator assembly have been shut down

due to an over-current, over-voltage , or under-voltage condition.

POWER UNLEVELED

There is either a hardware failure in the source or you have attempted to set

the power level too high. The analyzer allows the output power to be set

higher or lower than the speci ed available power range . However , these

output powers may be unleveled or unavailable . Check to see if the power

level you set is within speci cations . If it is , refer to the HP 8753D Network

Analyzer Service Guide for troubleshooting.

Show 8 more labels / short text on this page
Error Number · Error Number · Error Number · Error Number · Error Number · Error Number · Error Number · 10-18 Error Messages
Page 595 of the Agilent Technologies 8753D user manual — LOCAL
Page 595 · click to enlarge
Page 595

PRESS [MENU], SELECT CW (IF) FREQ, THEN SWEPT LO

When you are sweeping the RF and LO , the IF must be xed.

PRINT ABORTED

When you press the

LOCAL

key , the analyzer aborts output to the printer .

print color not supported with EPSON

Y ou have de ned the printer type as EPSON-P2. Color print is not supported

with this printer . The print will abort.

PRINTERbusy

The parallel port printer is not accepting data.

PRINTERerror

The parallel port printer is malfunctioning. The analyzer cannot complete the

copy function.

PRINTERnot connected

There is no printer connected to the parallel port.

PRINTERnot handshaking

The printer at the parallel port is not responding.

PRINTERnot on line

The printer at the parallel port is not set on line .

Show 9 more labels / short text on this page
Error Number · Error Number · Error Number · Error Number · Error Number · Error Number · Error Number · Error Number · Error Messages 10-19
Page 596 of the Agilent Technologies 8753D user manual — PROCESSING DISPLAY LIST
Page 596 · click to enlarge
Page 596
PRINTERnot on, not connected, wrong addrs

The printer does not respond to control. V erify power to the printer , and check

the HP-IB connection between the analyzer and the printer . Ensure that the

printer address recognized by the analyzer matches the HP-IB address set on

the printer itself.

PRINTERpaper error

There is a paper-related problem with the parallel port printer such as a paper

jam or out-of-paper condition.

PRINTERpower off

The power to the printer at the parallel port is o .

PRINT/PLOT IN PROGRESS, ABORT WITH LOCAL

If a print or plot is in progress and you attempt a second print or plot, this

message is displayed and the second attempt is ignored. T o abort a print or plot

in progress , press

LOCAL

.

PROBE POWER SHUT DOWN!

One or both of the probe power supplies have been shut down due to an

over-current, over-voltage , or under-voltage condition.

PROCESSING DISPLAY LIST

The display information is being processed for a screen print to a copy device

and stored in the copy spool bu er . During this time , the analyzer's resources

are dedicated to this task (which takes less than a few seconds).

Show 8 more labels / short text on this page
Error Number · Error Number · Error Number · Error Number · Error Number · Information · Message · 10-20 Error Messages
Page 597 of the Agilent Technologies 8753D user manual — SEQUENCE MAY HAVE CHANGED, CAN'T CONTINUE
Page 597 · click to enlarge
Page 597

PWR MTR NOT ON/CONNECTED OR WRONG ADDRS

The power meter cannot be accessed by the analyzer . V erify that the power

meter address and model number set in the analyzer match the address and

model number of the actual power meter .

REQUESTED DATA NOT CURRENTLY AVAILABLE

The analyzer does not currently contain the data you have requested. F or

example , this condition occurs when you request error term arrays and no

calibration is active .

SAVE FAILED. INSUFFICIENT MEMORY

Y ou cannot store an instrument state in an internal register due to insucient

memory . Increase the available memory by clearing one or more save/recall

registers and pressing

, or by storing les to a disk.

PRESET

SELECTED SEQUENCE IS EMPTY

The sequence you attempted to run does not contain instrument commands .

SELF TEST #n FAILED

Internal test #n has failed. Several internal test routines are executed at

instrument preset. The analyzer reports the rst failure detected. Refer to the

HP 8753D Network Analyzer Service Guide for troubleshooting information on

internal tests and the self-diagnose feature .

SEQUENCE ABORTED

The sequence running was stopped prematurely when you pressed the

LOCAL

key .

SEQUENCE MAY HAVE CHANGED, CAN'T CONTINUE

When you pause a sequence , you cannot continue it if you have modi ed it.

Y ou must start the sequence again.

Show 10 more labels / short text on this page
Error Number · Error Number · Error Number · Error Number · Service Error · Number 112 · Error Number · · Error Number · Error Messages 10-21
Page 598 of the Agilent Technologies 8753D user manual — SWEEP TIME INCREASED
Page 598 · click to enlarge
Page 598

SLIDES ABORTED (MEMORY REALLOCATION)

Y ou cannot perform sliding load measurements due to insucient memory .

Increase the available memory by clearing one or more save/recall registers and

, or by storing les to a disk.

PRESET

SOURCE POWER TURNED OFF, RESET UNDER POWER MENU

Y ou have exceeded the maximum power level at one of the inputs and power

has been automatically reduced. The annotation P+ indicates that power trip

has been activated. When this occurs , reset the power and then press

MENU

NNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

POWER

SOURCE PWR on OFF , to switch on the power .

STARTING COPY SPOOLER

The analyzer is beginning to output data from the spool bu er to the copy

device . The analyzer resumes normal operation; the data is being output to the

copy device in the background.

STOP/CW FREQ + OFFSET MUST BE < 3 GHz

The output frequency of the mixer cannot violate the minimum/maximum

frequency of the analyzer .

SWEEP MODE CHANGED TO CW TIME SWEEP

If you select external source auto or manual instrument mode and you do not

also select CW mode , the analyzer is automatically switched to CW .

SWEEP TIME INCREASED

Y ou have made instrument changes that cause the analyzer sweep time to be

automatically increased. Some parameter changes that cause an increase in

sweep time are narrower IF bandwidth, an increase in the number of points ,

and a change in sweep type .

Show 11 more labels / short text on this page
Error Number · pressing · Information · Message · · Information · Message · Error Number · Error Number · Error Number · 10-22 Error Messages
Page 599 of the Agilent Technologies 8753D user manual — THIS LIST FREQ INVALID IN HARM/3 GHZ RNG
Page 599 · click to enlarge
Page 599

SWEEP TIME TOO FAST

The fractional-N and digital IF circuits have lost synchronization. Refer to the

HP 8753D Network Analyzer Service Guide for troubleshooting information.

SWEEP TRIGGER SET TO HOLD

The instrument is in a hold state and is no longer sweeping. T o take a new

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

sweep , press

MENU

TRIGGER MENU

CONTINUOUS .

SYNTAX ERROR

Y ou have improperly formatted an HP-IB command. Refer to the HP 8753D

Network Analyzer Programmer's Guide for proper command syntax.

SYST CTRL OR PASS CTRL IN LOCAL MENU

The analyzer is in talker/listener mode . In this mode , the analyzer cannot

control a peripheral device on the bus . Use the local menu to change to system

controller or pass control mode .

TEST ABORTED

Y ou have prematurely stopped a service test.

THIS LIST FREQ INVALID IN HARM/3 GHZ RNG

Y ou have set frequencies in the list that are outside of the allowable frequency

range for harmonic measurements , or are greater than 3 GHz on instruments

without Option 006. Reduce the frequency range of the list.

Show 10 more labels / short text on this page
Error Number · Information · Message · · SINGLE or · Error Number · Error Number · Error Number · Error Number · Error Messages 10-23
Page 600 of the Agilent Technologies 8753D user manual — WAITING FOR DISK
Page 600 · click to enlarge
Page 600

TOO MANY NESTED SEQUENCES. SEQ ABORTED

Y ou can only nest sequences to a maximum level of six. The sequence will

abort if you nest more than six.

TOO MANY SEGMENTS OR POINTS

Y ou can have a maximum of 30 segments or 1632 points in frequency list mode .

In power meter calibrations , you can have a maximum of 12 segments for

power sensor cal factors and power loss functions .

TRANSFORM, GATE NOT ALLOWED

Y ou can perform a time domain transformation only in linear and CW sweep

types .

TROUBLE! CHECK SETUP AND START OVER

Y our equipment setup for the adjustment procedure in progress is not correct.

Check the setup diagram and instructions HP 8753D Network Analyzer Service

Guide. Start the procedure again.

WAITING FOR CLEAN SWEEP

In single sweep mode , the instrument ensures that all changes to the

instrument state , if any , have been implemented before taking the sweep . The

command that you have initiated is being processed and will not be complete

until the new sweep is completed. An asterisk * is displayed in the left margin

until a complete fresh sweep has been taken.

WAITING FOR DISK

This message is displayed between the start and nish of a read or write

operation to a disk.

Show 10 more labels / short text on this page
Error Number · Error Number · Error Number · Service Error · Number 115 · Information · Message · Information · Message · 10-24 Error Messages
Page 601 of the Agilent Technologies 8753D user manual — WRONG DISK FORMAT, INITIALIZE DISK
Page 601 · click to enlarge
Page 601

WAITING FOR HP-IB CONTROL

Y ou have instructed the analyzer to use pass control (USEP ASC). When you

send the analyzer an instruction that requires active controller mode , the

analyzer requests control of the bus and simultaneously displays this message .

If the message remains , the system controller is not relinquishing the bus .

WRITE ATTEMPTED WITHOUT SELECTING INPUT TYPE

Y ou have sent the data header \#A" to the analyzer with no preceding input

command (such as INPUD A T A). The instrument recognized the header but did

not know what type of data to receive . Refer to the HP 8753D Network

Analyzer Programmer's Guide for command syntax information.

WRONG DISK FORMAT, INITIALIZE DISK

Y ou have attempted to store , load, or read le titles , but your disk format does

not conform to the Logical Interchange F ormat (LIF) or DOS format. Y ou must

initialize the disk before reading or writing to it.

Show 5 more labels / short text on this page
Information · Message · Error Number · Error Number · Error Messages 10-25
Page 602 of the Agilent Technologies 8753D user manual — 25 PRINT ABORTED
Page 602 · click to enlarge
Page 602

Error Messages in Numerical Order

Refer to the alphabetical listing for explanations and suggestions for solving the problems .

1 OPTIONAL FUNCTION; NOT INST ALLED

2 INV ALID KEY

3 CORRECTION CONST ANTS NOT STORED

4 PHASE LOCK CAL F AILED

5 NO IF FOUND: CHECK R INPUT LEVEL

6 POSSIBLE F ALSE LOCK

7 NO PHASE LOCK: CHECK R INPUT LEVEL

8 PHASE LOCK LOST

9 LIST T ABLE EMPTY

10 CONTINUOUS SWITCHING NOT ALLOWED

11 SWEEP TIME INCREASED

12 SWEEP TIME TOO F AST

13 A VERA GING INV ALID ON NON-RA TIO MEASURE

14 FUNCTION NOT V ALID

15 NO MARKER DEL T A - SP AN NOT SET

16 TRANSFORM, GA TE NOT ALLOWED

17 DEMODULA TION NOT V ALID

19 LIST T ABLE EMPTYoccurs if user selects LIST sweep type but there is no

list freq. table

20 AIR FLOW RESTRICTEDCHECK F AN FIL TER

21 POWER SUPPL Y HOT!

22 POWER SUPPL Y SHUT DOWN!

23 PROBE POWER SHUT DOWN!

24 PRINTERnot on, not connect, wrong addrs

25 PRINT ABORTED

26 PLOTTERnot on, not connect, wrong addrs
Show 5 more labels / short text on this page
Error · Error · Number · 18 not used · 10-26 Error Messages
Page 603 of the Agilent Technologies 8753D user manual — 56 INSTRUMENT ST A TE MEMORY CLEARED
Page 603 · click to enlarge
Page 603

27 PLOT ABORTED

28 PLOTTER NOT READY-PINCH WHEELS UP

30 REQUESTED D A T A NOT CURRENTL Y A V AILABLE

31 ADDRESSED TO T ALK WITH NOTHING TO SA Y

32 WRITE A TTEMPTED WITHOUT SELECTING INPUT TYPE

33 SYNT AX ERROR

34 BLOCK INPUT ERROR

35 BLOCK INPUT LENGTH ERROR

36 SYST CTRL OR P ASS CTRL IN LOCAL MENU

37 ANOTHER SYSTEM CONTROLLER ON HP-IB BUS

38 DISKnot on, not connected, wrong addrs

39 DISK HARDW ARE PROBLEM

40 DISK MEDIUM NOT INITIALIZED

41 NO DISK MEDIUM IN DRIVE

42 FIRST CHARA CTER MUST BE A LETTER

43 ONL Y LETTERS AND NUMBERS ARE ALLOWED

44 NOT ENOUGH SP A CE ON DISK FOR STORE

45 NO FILE(S) FOUND ON DISK

46 ILLEGAL UNIT OR V OLUME NUMBER

47 INITIALIZA TION F AILED

48 DISK IS WRITE PROTECTED

49 DISK WEAR-REPLA CE DISK SOON

50 TOO MANY SEGMENTS OR POINTS

51 INSUFFICIENT MEMORY

54 NO V ALID MEMORY TRA CE

55 NO V ALID ST A TE IN REGISTER

56 INSTRUMENT ST A TE MEMORY CLEARED

Show 5 more labels / short text on this page
Error · Error · Number · 52 not used · Error Messages 10-27
Page 604 of the Agilent Technologies 8753D user manual — 124 SELECTED SEQUENCE IS EMPTY
Page 604 · click to enlarge
Page 604

57 O VERLO AD ON INPUT R, POWER REDUCED

58 O VERLO AD ON INPUT A, POWER REDUCED

59 O VERLO AD ON INPUT B , POWER REDUCED

61 HP 8753 SOURCE P ARAMETERS CHANGED

63 CALIBRA TION REQUIRED

64 CURRENT P ARAMETER NOT IN CAL SET

65 CORRECTION AND DOMAIN RESET

66 CORRECTION TURNED OFF

67 DOMAIN RESET

68 ADDITIONAL ST AND ARDS NEEDED

69 NO CALIBRA TION CURRENTL Y IN PROGRESS

70 NO SP A CE FOR NEW CAL. CLEAR REGISTERS

71 MORE SLIDES NEEDED

72 EXCEEDED 7 ST AND ARDS PER CLASS

73 SLIDES ABORTED (MEMORY REALLOCA TION)

74 CALIBRA TION ABORTED

75 FORMA T NOT V ALID FOR MEASUREMENT

77 WRONG DISK FORMA T , INITIALIZE DISK

111 DEADLOCK

112 SELF TEST #n F AILED

113 TEST ABORTED

114 NO F AIL FOUND

115 TROUBLE! CHECK SETUP AND ST ART O VER

116 POWER METER INV ALID

117 PWR MTR: NOT ON/CONNECTED OR WRONG ADDRS

118 POWER METER NOT SETTLED

119 DEVICEnot on, not connect, wrong addrs

123 NO MEMORY A V AILABLE FOR INTERPOLA TION

124 SELECTED SEQUENCE IS EMPTY

Show 5 more labels / short text on this page
Error · Error · Number · 60 not used · 10-28 Error Messages
Page 605 of the Agilent Technologies 8753D user manual — 168 INSUFFICIENT MEMORY FOR PRINT/PLOT
Page 605 · click to enlarge
Page 605

125 DUPLICA TING TO THIS SEQUENCE NOT ALLOWED

126 NO MEMORY A V AILABLE FOR SEQUENCING

127 CAN'T STORE/LO AD SEQUENCE, INSUFFICIENT MEMORY

130 D2/D1 INV ALID WITH SINGLE CHANNEL

131 FUNCTION NOT V ALID DURING MOD SEQUENCE

132 MEMORY FOR CURRENT SEQUENCE IS FULL

133 THIS LIST FREQ INV ALID IN HARM/3 GHZ RNG

140 FREQ OFFSET ONL Y V ALID IN NETWORK ANAL YZER MODE

141 STOP/CW FREQ + OFFSET MUST BE < 3 GHz

144 NO LIMIT LINES DISPLA YED

148 EXTERNAL SOURCE MODE REQUIRES CW TIME

150 LOG SWEEP REQUIRES 2 OCT A VE MINIMUM SP AN

151 SA VE F AILED / INSUFFICIENT MEMORY

152 D2/D1 INV ALIDCH1 CH2 NUM PTS DIFFERENT

153 SEQUENCE MA Y HA VE CHANGED , CAN'T CONTINUE

154 INSUFFICIENT MEMORY , PWR MTR CAL OFF

157 SEQUENCE ABORTED

159 CH1 (CH2) T ARGET V ALUE NOT FOUND

161 PRESS [MENU], SELECT CW (IF) FREQ, THEN SWEPT LO

162 EXT SRC: NOT ON/CONNECTED OR WRONG ADDR

163 FUNCTION ONL Y V ALID DURING MOD SEQUENCE

164 TOO MANY NESTED SEQUENCES. SEQ ABORTED

165 P ARALLEL PORT NOT A V AILABLE FOR GPIO

166 PRINT/PLOT IN PROGRESS, ABORT WITH LOCAL

167 P ARALLEL PORT NOT A V AILABLE FOR COPY

168 INSUFFICIENT MEMORY FOR PRINT/PLOT

169 HPIB COPY IN PROGRESS, ABORT WITH LOCAL

170 COPY :device not responding; copy aborted

Show 5 more labels / short text on this page
Error · Error · Number · 145 not used · Error Messages 10-29
Page 606 of the Agilent Technologies 8753D user manual — 187 SWEEP MODE CHANGED TO CW TIME SWEEP
Page 606 · click to enlarge
Page 606
171 PRINTERpaper error
172 PRINTERnot on line
173 PRINTERnot connected
174 PRINTERpower o
175 PRINTERerror
176 PRINTERbusy
177 PRINTERnot handshaking

178 print color not supported with EPSON

179 POWER UNLEVELED

180 DOS NAME LIMITED TO 8 CHARS + 3 CHAR EXTENSION

181 B AD FREQ FOR HARMONIC OR FREQ OFFSET

182 LIST MODE OFFINV ALID WITH LO FREQ

183 B A TTERY F AILED . ST A TE MEMORY CLEARED

184 B A TTERY LOW! STORE SA VE REGS TO DISK

185 CANNOT FORMA T DOS DISKS ON THIS DRIVE

187 SWEEP MODE CHANGED TO CW TIME SWEEP

188 DIRECTORY FULL

189 DISK READ/WRITE ERROR

190 DISK MESSA GE LENGTH ERROR

191 EXT SOURCE NOT READY FOR TRIGGER

192 FILE NOT FOUND

193 ASCIIMISSING 'BEGIN' statement
194 ASCIIMISSING 'CITIFILE' statement
195 ASCIIMISSING 'D A T A' statement
196 ASCIIMISSING 'V AR' statement
Show 4 more labels / short text on this page
Error · Error · Number · 10-30 Error Messages
Page 607 of the Agilent Technologies 8753D user manual — 197 FILE NOT FOUND OR WRONG TYPE
Page 607 · click to enlarge
Page 607

197 FILE NOT FOUND OR WRONG TYPE

198 NOT ALLOWED DURING POWER METER CAL

199 CANNOT MODIFY F A CTORY PRESET

200 ALL REGISTERS HA VE BEEN USED

201 FUNCTION NOT V ALID FOR INTERNAL MEMORY

202 FUNCTION NOT A V AILABLE

203 CANNOT READ/WRITE HFS FILE SYSTEM

204 FREQS CANNOT BE CHANGED , TOO MANY POINTS

Show 4 more labels / short text on this page
Error · Error · Number · Error Messages 10-31
Page 608 of the Agilent Technologies 8753D user manual — Chapter 10 Error Messages
Page 608 · click to enlarge
Page 608
Page 608 of the Agilent Technologies 8753D user manual — Chapter 10 Error Messages
Page 609 of the Agilent Technologies 8753D user manual — Chapter 11 Compatible Peripherals
Page 609 · click to enlarge
Page 609

This chapter contains the following information:

Connecting and con guring peripherals

Where to Look for More Information

A dditional information about many of the topics discussed in this chapter is located in the

following areas:

Chapter 2, \Making Measurements ," contains step-by-step procedures for making

measurements or using particular functions .

Chapter 6, \Application and Operation Concepts ," contains explanatory-style information

about many applications and analyzer operation.

Measurement A ccessories A vailable

The following calibration kits contain precision standards and required adapters of the

indicated connector type . The standards (known devices) facilitate measurement calibration,

also called vector error correction. Refer to the data sheet and ordering guide for additional

information. P arts numbers for the standards are in their manuals .

HP 85031B 7 mm Calibration Kit

HP 85032B 50 Ohm Type-N Calibration Kit

HP 85033D 3.5 mm Calibration Kit

HP 85033C 3.5 mm Calibration Kit

HP 85036B 75 Ohm Type-N Calibration Kit

HP 85039A 75 Ohm Type-F Calibration Kit

The HP 85033D is the recommended 3.5 mm calibration kit as it provides

greater measurement accuracy than the HP 85033C and it is easier to use due

to one-piece opens .

Compatible P eripherals 11-1

Show 5 more labels / short text on this page
Compatible P eripherals · Measurement accessories available · System accessories available · HP-IB programming overview · Calibration Kits
Page 610 of the Agilent Technologies 8753D user manual — Chapter 11 Compatible Peripherals
Page 610 · click to enlarge
Page 610

V eri cation Kit

A ccurate operation of the analyzer system can be veri ed by measuring known devices other

than the standards used in calibration, and comparing the results with recorded data.

HP 85029B 7 mm V eri cation Kit

This kit contains traceable precision 7 mm devices used to con rm the system's error-corrected

measurement uncertainty performance . Also included is veri cation data on a 3.5 inch disk,

together with a hard-copy listing. A system veri cation procedure is provided with this kit and

also in the HP 8753D Service Guide.

T est P ort Return Cables

The following RF cables are used to connect a two-port device between the test ports . These

cables provide shielding for high dynamic range measurements .

HP 11857D 7 mm T est P ort Return Cable Set

This set consists of a pair of test port return cables that can be used in measurements of 7 mm

devices . They can also be used with connectors other than 7 mm by using the appropriate

precision adapters .

HP 11857B 75 Ohm Type-N T est P ort Return Cable Set

This set consists of test port return cables for use with the HP 8753D Option 075.

HP 11852B 50 to 75 Ohm Minimum Loss P ad.

This device converts impedance from 50 ohms to 75 ohms or from 75 ohms to 50 ohms . It is

used to provide a low SWR impedance match between a 75 ohm device under test and the

HP 8753D network analyzer (without Option 075).

These adapter kits contain the connection hardware required for making measurements on

devices of the indicated connector type .

HP 11853A 50 Ohm Type-N A dapter Kit

HP 11854A 50 Ohm BNC A dapter Kit

HP 11855A 75 Ohm Type-N A dapter Kit

HP 11856A 75 Ohm BNC A dapter Kit

11-2 Compatible P eripherals

A dapter Kits

Page 611 of the Agilent Technologies 8753D user manual — Chapter 11 Compatible Peripherals
Page 611 · click to enlarge
Page 611

Transistor T est Fixtures

The following Hewlett-P ackard transistor test xtures are compatible with the HP 8753D .

A dditional test xtures for transistors and other devices are available from Inter-Continental

Microwave . T o order their catalog, request HP literature number 5091{4254E. Or contact

Inter{Continental Microwave as follows:

Santa Clara, CA 95054-1524

HP 11600B and 11602B Transistor Fixtures .

These xtures are used to hold devices for S-parameter measurements in a 50 ohm coaxial

circuit. They can be used to measure bipolar or eld-e ect transistors in several con gurations ,

from dc to 2.0 GHz. The HP 11600B accepts transistors with TO-18 to TO-72 package

dimensions , and the HP 11602B accepts transistors with TO-5 to TO-12 package dimensions .

Both xtures can also be used to measure other circuit elements such as diodes , resistors , or

inductors , which have 0.016 to 0.019 inch diameter leads .

HP 11608A Option 003 Transistor Fixture .

This xture is designed to be user-milled to hold stripline transistors for S-parameter

measurements . Option 003 is pre-milled for 0.205 inch diameter disc packages , such as the

HP HP A C-200.

HP 11858A Transistor Fixture A dapter .

This adapter provides a rigid RF cable interconnection between the HP 8753D test port

connectors and the HP 11600B , 11602B , or 11608A transistor xture .

Compatible P eripherals 11-3

1515 Wyatt Drive · (tel) 408 727-1596 · (fax) 408 727-0105

Page 612 of the Agilent Technologies 8753D user manual — Chapter 11 Compatible Peripherals
Page 612 · click to enlarge
Page 612

System A ccessories A vailable

The HP 85043D system cabinet is designed to rack mount the analyzer in a system

con guration. The 132 cm (52 in) system cabinet includes a bookcase , a drawer , and a

convenient work surface .

The HP 1181A system testmobile is designed to provide mobility for instruments , test systems ,

and work stations . This system testmobile can hold units up to 610-mm (24 in) deep and has a

load capacity of up to 90 kg (200 lb) on the tilt tray and 227 kg (500 lb) total.

The analyzer is capable of plotting or printing displayed measurement results directly (without

the use of an external computer) to a compatible peripheral. The analyzer supports HP-IB ,

serial, and parallel peripherals . Most Hewlett-P ackard desktop printers and plotters are

compatible with the analyzer . Some common compatible peripherals are listed here (some are

no longer available for purchase but are listed here for your reference):

These plotters are compatible:

HP 7440A ColorPro Eight-P en Color Graphics Plotter

HP 7470A Two-P en Graphics Plotter

HP 7475A Six-P en Graphics Plotter

HP 7550A/B High-Speed Eight-P en Graphics Plotter

These printers are compatible:

HP C2621A, DeskJet P ortable 310

HP C2634A, DeskJet P ortable 320

HP C2655A, DeskJet P ortable 340

HP C2642A, DeskJet 400

HP C2106A, DeskJet 500

HP C2114A, DeskJet 500C

HP C2170A, DeskJet 520

HP C2162A, DeskJet 540

HP C2121A, DeskJet 550C

HP C2168A, DeskJet 560C

HP C2184A, DeskJet 600

HP C2164A, DeskJet 660C

HP C4567A, DeskJet 682C

HP C4562A, DeskJet 690C

HP C2145A, DeskJet 850C

11-4 Compatible P eripherals

System Cabinet · System T estmobile · Plotters and Printers

Page 613 of the Agilent Technologies 8753D user manual — Chapter 11 Compatible Peripherals
Page 613 · click to enlarge
Page 613

HP C1676A, DeskJet 1200C (can also be used to plot)

HP C3540A, DeskJet 1600C (can also be used to plot)

All LaserJets (can also be used to plot)

HP 3630A P aintJet Color Graphics Printer

Epson printers which are compatible with the Epson ESC/P2 printer control language , such

as the LQ570, are also supported by the analyzer . Older Epson printers , however , such as the

FX-80, will not work with the analyzer .

The analyzer has the capability of storing instrument states directly to its internal memory ,

to an internal disk drive , or to an external disk drive . The internal 3.5 inch oppy disk can

be initialized in both LIF and DOS formats and is capable of reading and writing data in both

formats . Using the internal disk drive is the preferred method, but the capability to use

external disk drives still exists . Most external disks using CS80 protocol are compatible .

The analyzer does not support the LIF-1 (hierarchy le system) directory

format.

Do not use the older single-sided disks in the analyzer's internal drive .

An HP-IB cable is required for interfacing the analyzer with a plotter , printer , external disk

drive , or computer . The cables available are:

HP 10833A HP-IB Cable , 1.0 m (3.3 ft.)

HP 10833B HP-IB Cable , 2.0 m (6.6 ft.)

HP 10833D HP-IB Cable , 0.5 m (1.6 ft.)

HP C2912B Centronics (P arallel) Interface Cable , 3.0 m (9.9 ft.)

HP C2913A RS-232C Interface Cable , 1.2 m (3.9 ft.)

HP C2914A Serial Interface Cable , 1.2 m (3.9 ft.)

HP 24542G Serial Interface Cable , 3 m (9.9 ft.)

HP 24542D P arallel Interface Cable , 2 m (6 ft.)

HP 92284A P arallel Interface Cable , 2 m (6 ft.)

Compatible P eripherals 11-5

Show 5 more labels / short text on this page
HP 2227B QuietJet · HP 2225A ThinkJet · Mass Storage · HP-IB Cables · Interface Cables
Page 614 of the Agilent Technologies 8753D user manual — Chapter 11 Compatible Peripherals
Page 614 · click to enlarge
Page 614

A keyboard can be connected to the analyzer for data input such as titling les .

A keyboard can be connected to the analyzer for data input such as titling les . The

HP C1405A Option AB A keyboard with the HP part number C1405-60015 adapter is suitable for

this purpose . Or , the analyzer is designed to accept most PC-A T-compatible keyboards with a

standard DIN connector . Keyboards with a mini-DIN connector are compatible with the HP part

number C1405-60015 adapter .

An external controller is not required for measurement calibration or time domain capability .

However , some performance test procedures are semi-automated and require the use of an

external controller . (The system veri cation procedure does not require an external controller .)

The system can be automated with the following computers:

IBM PC compatible computer with an HP-IB/GP-IB interface card

HP 9000 series 200/300 workstation

HP 9000 series 700 workstation with HP B ASIC-UX

F or more information about compatible computers , call your nearest Hewlett-P ackard Sales and

Service oce .

A set of sample measurement programs is provided with the HP 8753D Network Analyzer

Programmer's Guide on a 3.5 inch disk. The programs include typical measurements to be

used as an introductory example for programming the analyzer over HP-IB . It is designed to be

easily modi ed for use in developing programs for speci c needs . The programs are compatible

with HP B ASIC versions 2.0 and later , QuickBasic , and QuickC, and will run on an IBM PC

compatible computer (as well as the HP 9000 series 200/300/700 workstations), using any HP

8753D compatible printer or plotter .

The following additional software products are also available for the HP 8753D:

HP 85190A IC-CAP Modelling Suite

(MDS) HP Microwave Design System

The analyzer can drive both its internal display and an external monitor simultaneously . One

recommended color monitor is the HP 35741A/B .

External Monitor Requirements:

60 Hz vertical refresh rate

25.5 kHz horizontal refresh rate

RGB with synchronization on green

75 ohm video input impedance

video amplitude 1 Vp-p (0.7 V = white , 0 V = black, 0.3 V = synchronization)

11-6 Compatible P eripherals

Show 6 more labels / short text on this page
K eyboards · Controller · Sample Software · T ouchstone · EEsof Series-IV · External Monitors
Page 615 of the Agilent Technologies 8753D user manual — HP-IB HP 10833A/B/D
Page 615 · click to enlarge
Page 615

Connecting the P eripheral Device

Connect the peripheral to the corresponding interface port.

Printer Interface Recommended Cable

P arallel HP 92284A

HP-IB HP 10833A/B/D

Figure 11-1. P eripheral Connections to the Analyzer

The keyboard can be connected to the analyzer while the power is on or o .

Compatible P eripherals 11-7

Connecting P eripherals · Serial HP 24542G

Page 616 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 616 · click to enlarge
Page 616

Con guring the Analyzer for the P eripheral

All copy con guration settings are stored in non-volatile memory . Therefore , they are not

a ected if you press

or switch o the analyzer power .

PRESET

If the P eripheral is a Printer

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SET ADDRESSES

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRINTER PORT

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRNTR TYPE until the correct printer choice

LOCAL

appears:

NNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNN

DeskJet (except for HP DeskJet 540 and DeskJet 850C)

NNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNN

Epson-P2 (printers that conform to the ESC/P2 printer control language)

NNNNNNNNNNNNNNNNNNNN

DJ 540 (for use with the HP DeskJet 540 and DeskJet 850C)

NNNNNNNNNNNNNNNNNNNN

DJ 540 converts 100 dpi raster information to 300 dpi raster format.

If your DeskJet printer does not support the 100 dpi raster format and your

NNNNNNNNNNNNNNNNNNNN

printing results seem to be less than normal size , select

DJ 540 .

2. Select one of the following printer interfaces:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRNTR PORT HPIB if your printer has an HP-IB interface , and then con gure the

print function as follows:

a. Enter the HP-IB address of the printer , followed by

.

b.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOCAL

SYSTEM CONTROLLER if there is no external controller connected to

the HP-IB bus .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

c. Press

LOCAL

USE PASS CONTROL if there is an external controller connected to

the HP-IB bus .

NNNNNNNNNNNNNNNNNNNNNNNNNN

PARALLEL if your printer has a parallel (centronics) interface , and then con gure

the print function as follows:

and then select the parallel port interface function by pressing

NNNNNNNNNNNNNNNNNNNNNNNNNN

PARALLEL

LOCAL

until the correct function appears:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PARALLEL [COPY] , the parallel port is dedicated for normal copy

device use (printers or plotters).

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PARALLEL [GPIO] , the parallel port is dedicated for general purpose

I/O , and cannot be used for printing or plotting.

11-8 Compatible P eripherals

Show 14 more labels / short text on this page
Press · · ThinkJet (QuietJet) · LaserJet · PaintJet · Selecting · Choose · Press · and · and · Choose · Press · If you choose · If you choose
Page 617 of the Agilent Technologies 8753D user manual — LOCAL
Page 617 · click to enlarge
Page 617

NNNNNNNNNNNNNNNNNNNN

SERIAL if your printer has a serial (RS-232) interface , and then con gure the print

function as follows:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRINTER BAUD RATE and enter the printer's baud rate , followed by

.

a. Press

b. T o select the transmission control method that is compatible with your printer , press

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

XMIT CNTRL (transmit control - handshaking protocol) until the correct method

appears:

NNNNNNNNNNNNNNNNNNNNNNNNNN

Xon-Xoff , the handshake method allows the printer to control the

data exchange by transmitting control characters to the network analyzer .

NNNNNNNNNNNNNNNNNNNNNNN

DTR-DSR , the handshake method allows the printer to control the data

exchange by setting the electrical voltage on one line of the RS-232 serial cable .

NNNNNNNNNNNNNNNNNNNNNNN

DTR-DSR handshake takes place in the hardware rather than the

rmware or software , it is the fastest transmission control method.

If the P eripheral is a Plotter

HPGL/2 Compatible Printer (used as a plotter)

N

SET ADDRESSES

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRINTER PORT and then press

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRNTR TYPE until the correct

LOCAL

printer choice appears:

NNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNN

DeskJet (only DeskJet 1200C and DeskJet 1600C)

NNNNNNNNNNNNNNNNNNNNNNNNNN

LaserJet (only LaserJet III and IV)

NNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNN

Epson-P2 (printers that conform to the ESC/P2 printer control language)

2. Con gure the analyzer for one of the following printer interfaces:

N

PRNTR PORT HPIB if your printer has an HP-IB interface , and then con gure the

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

print function as follows:

a. Enter the HP-IB address of the printer (default is 01), followed by

.

b.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOCAL

SYSTEM CONTROLLER if there is no external controller connected to

the HP-IB bus .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

c. Press

LOCAL

USE PASS CONTROL if there is an external controller connected to

the HP-IB bus .

Compatible P eripherals 11-9

Show 12 more labels / short text on this page
Choose · If you choose · If you choose · Because the · Press · · ThinkJet (QuietJet) · PaintJet · Choose · Press · and · and
Page 618 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 618 · click to enlarge
Page 618

NNNNNNNNNNNNNNNNNNNNNNNNNN

PARALLEL if your printer has a parallel (centronics) interface , and then con gure

the print function as follows:

NNNNNNNNNNNNNNNNNNNNNNNNNN

LOCAL

and then select the parallel port interface function by pressing

PARALLEL

until the correct function appears:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PARALLEL [COPY] , the parallel port is dedicated for normal copy

device use (printers or plotters).

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PARALLEL [GPIO] , the parallel port is dedicated for general purpose

I/O , and cannot be used for printing or plotting.

N

NNNNNNNNNNNNNNNNNNN

SERIAL if your printer has a serial (RS-232) interface , and then con gure the print

function as follows:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

a. Press

PRINTER BAUD RATE and enter the printer's baud rate , followed by

.

b. T o select the transmission control method that is compatible with your printer , press

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

XMIT CNTRL (transmit control - handshaking protocol) until the correct method

appears:

N

NNNNNNNNNNNNNNNNNNNNNNNNN

Xon-Xoff , the handshake method allows the printer to control the

data exchange by transmitting control characters to the network analyzer .

NNNNNNNNNNNNNNNNNNNNNNN

DTR-DSR , the handshake method allows the printer to control the data

exchange by setting the electrical voltage on one line of the RS-232 serial cable .

N

NNNNNNNNNNNNNNNNNNNNNN

DTR-DSR handshake takes place in the hardware rather than the

rmware or software , it is the fastest transmission control method.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

3. Press

LOCAL

SET ADDRESSES

PLOTTER PORT and then

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLTR TYPE [HPGL PRT] appears .

11-10 Compatible P eripherals

Show 10 more labels / short text on this page
Choose · Press · If you choose · If you choose · Choose · If you choose · If you choose · Because the · · PLTR TYPE until
Page 619 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNN
Page 619 · click to enlarge
Page 619

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

1. Press

LOCAL

SET ADDRESSES

PLOTTER PORT and then

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLTR TYPE [PLOTTER] appears .

2. Con gure the analyzer for one of the following plotter interfaces:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLTR PORT HPIB if your plotter has an HP-IB interface , and then con gure the

plot function as follows:

a. Enter the HP-IB address of the printer (default is 05), followed by

.

b.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOCAL

SYSTEM CONTROLLER if there is no external controller connected to

the HP-IB bus .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

c. Press

LOCAL

USE PASS CONTROL if there is an external controller connected to

the HP-IB bus .

NNNNNNNNNNNNNNNNNNNNNNNNNN

PARALLEL if your printer has a parallel (centronics) interface , and then con gure

the print function as follows:

and then select the parallel port interface function by pressing

NNNNNNNNNNNNNNNNNNNNNNNNNN

PARALLEL

LOCAL

until the correct function appears:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PARALLEL [COPY] , the parallel port is dedicated for normal copy

device use (printers or plotters).

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PARALLEL [GPIO] , the parallel port is dedicated for general purpose

I/O , and cannot be used for printing or plotting.

NNNNNNNNNNNNNNNNNNNN

SERIAL if your printer has a serial (RS-232) interface , and then con gure the print

function as follows:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRINTER BAUD RATE and enter the printer's baud rate , followed by

.

a. Press

b. T o select the transmission control method that is compatible with your printer , press

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

XMIT CNTRL (transmit control - handshaking protocol) until the correct method

appears:

NNNNNNNNNNNNNNNNNNNNNNNNNN

Xon-Xoff , the handshake method allows the printer to control the

data exchange by transmitting control characters to the network analyzer .

NNNNNNNNNNNNNNNNNNNNNNN

DTR-DSR , the handshake method allows the printer to control the data

exchange by setting the electrical voltage on one line of the RS-232 serial cable .

NNNNNNNNNNNNNNNNNNNNNNN

DTR-DSR handshake takes place in the hardware rather than the

rmware or software , it is the fastest transmission control method.

Compatible P eripherals 11-11

Show 15 more labels / short text on this page
P en Plotter · · PLTR TYPE until · Choose · Press · and · and · Choose · Press · If you choose · If you choose · Choose · If you choose · If you choose · Because the
Page 620 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 620 · click to enlarge
Page 620

If the P eripheral is a P ower Meter

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SET ADDRESSES .

LOCAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

POWER MTR[ ] until the correct selection appears:

HP 436A

HP 437B or 438A

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

3. Press
ADDRESSP MTR/HP-IB and con gure the power meter as follows:

a. Enter the HP-IB address of the power meter , followed by

.

b.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOCAL

SYSTEM CONTROLLER if there is no external controller connected to the

HP-IB bus .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

c. Press

LOCAL

USE PASS CONTROL if there is an external controller connected to the

HP-IB bus .

If the P eripheral is an External Disk Drive

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

DISK UNIT NUMBER and enter the drive where your disk is located, followed

LOCAL

.

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

If your storage disk is partitioned, press

VOLUME NUMBER and enter the volume number

where you want to store the instrument state le .

N

SET ADDRESSES

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ADDRESS: DISK .

4. Press

LOCAL

and select one of the following:

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SYSTEM CONTROLLER if there is no external controller connected.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PASS CONTROL when an external controller is connected and the analyzer needs

to control peripherals directly . This mode causes the analyzer to request control from the

external controller whenever it needs to communicate with a peripheral.

If the P eripheral is a Computer Controller

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LOCAL

SET ADDRESSES

ADDRESS: CONTROLLER .

2. Press

and select one of the following:

LOCAL

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TALKER/LISTENER to allow the computer controller to be involved in all

peripheral access operations .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PASS CONTROL when an external controller is connected and the analyzer needs

to control peripherals directly . This mode causes the analyzer to request control from the

external controller whenever it needs to communicate with a peripheral.

11-12 Compatible P eripherals

Show 16 more labels / short text on this page
Press · · Press · Press · and · and · Press · · by · Press · Choose · Choose · Press · · Choose · Choose
Page 621 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 621 · click to enlarge
Page 621

Con guring the Analyzer to Produce a Time Stamp

Y ou can set a clock, and then activate it, if you want the time and date to appear on your

hardcopies .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SYSTEM

SET CLOCK .

NNNNNNNNNNNNNNNNNNNNNNNNNN

SET YEAR and enter the current year , followed by

.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SET MONTH and enter the current month of the year , followed

.

NNNNNNNNNNNNNNNNNNNNNNN

SET DAY and enter the current day of the month, followed by

.

NNNNNNNNNNNNNNNNNNNNNNNNNN

SET HOUR and enter the current hour of the day (0-23), followed by

.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SET MINUTES and enter the next immediate minute , followed by

.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

ROUND SECONDS when the current time is exactly as you have set it.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

TIME STAMP ON appears on the softkey label.

Compatible P eripherals 11-13

Show 10 more labels / short text on this page
Press · · Press · Press · Press · Press · Press · Press · Press · TIME STAMP until
Page 622 of the Agilent Technologies 8753D user manual — Chapter 11 Compatible Peripherals
Page 622 · click to enlarge
Page 622

The analyzer is factory-equipped with a remote programming digital interface using the

Hewlett-P ackard Interface Bus (HP-IB). HP-IB is Hewlett-P ackard's hardware , software ,

documentation, and support for IEEE 488.1 and IEC-625, worldwide standards for interfacing

instruments . The HP-IB lets you control the analyzer with an external computer that sends

commands or instructions to and receives data from the analyzer . This provides a remote

operator with the same control of the instrument available to the local operator , except for

control of the power line switch and some internal tests .

In addition, without the use of an external computer , the analyzer can use HP-IB to output

measurement results directly to a compatible printer or plotter and to store data to an external

disk drive . It can also control a power meter for power calibration and, through a subset of

HP-GL (Hewlett-P ackard Graphics Language), user graphics can be plotted on the analyzer

display .

F or more complete information on programming the analyzer remotely over HP-IB , refer to

the HP 8753D Network Analyzer Programmer's Guide.

F or a complete general description of the HP-IB , refer to the Tutorial Description of the

Hewlett-P ackard Interface Bus, HP publication 5952-0156 and to Condensed Description of

the Hewlett-P ackard Interface Bus (HP part number 59401-90030).

F or more information on the IEEE-488.1 standard, refer to IEEE Standard Digital Interface

for Programmable Instrumentation, published by the Institute of Electrical and Electronics

Engineers , Inc ., 345 East 47th Street, New Y ork, New Y ork 10017.

11-14 Compatible P eripherals

HP-IB Programming Overview

Page 623 of the Agilent Technologies 8753D user manual — Chapter 11 Compatible Peripherals
Page 623 · click to enlarge
Page 623

The Hewlett-P ackard Interface Bus (HP-IB) is Hewlett-P ackard's hardware , software ,

documentation, and support for IEEE 488.2 and IEC-625 worldwide standards for interfacing

instruments . This interface allows you to operate the analyzer and peripherals in two methods:

by an external system controller

by the network analyzer in system-controller mode

The HP-IB employs a party-line bus structure in which up to 15 devices can be connected

on one contiguous bus . The interface consists of 16 signal lines and 8 ground lines within a

shielded cable . With this cabling system, many di erent types of devices including instruments ,

computers , power meters , plotters , printers , and disk drives can be connected in parallel.

Every HP-IB device must be capable of performing one or more of the following interface

functions:

A talker is a device capable of transmitting device-dependent data when addressed to talk.

There can be only one active talker at any given time . Examples of this type of device include:

The network analyzer is a talker when it sends trace data or marker information over the bus .

A listener is a device capable of receiving device-dependent data over the interface when

addressed to listen. There can be as many as 14 listeners connected to the interface at any

given time . Examples of this type of device include:

The network analyzer is a listener when it is controlled over the bus by a system controller .

A controller is de ned as a device capable of:

1. managing the operation of the bus
2. addressing talkers and listeners

There can be only one active controller on the interface at any time . Examples of controllers

include desktop computers , minicomputers , workstations , and the network analyzer . In a

multiple-controller system, active control can be passed between controllers , but there can

only be one system controller connected to the interface . The system controller acts as the

master and can regain active control at any time . The analyzer is an active controller when it

plots , prints , or stores to an external disk drive in the pass-control mode . The analyzer is also a

system controller when it is operating in the system-controller mode .

Compatible P eripherals 11-15

Show 13 more labels / short text on this page
HP-IB Operation · Device Types · T alker · power meters · disk drives · voltmeters · counters · tape readers · Listener · printers · power supplies · signal generators · Controller
Page 624 of the Agilent Technologies 8753D user manual — Chapter 11 Compatible Peripherals
Page 624 · click to enlarge
Page 624
Agilent Technologies 8753D — Chapter 11 Compatible Peripherals diagram (page 624)
Fig. 259 · Chapter 11 Compatible Peripherals

Figure 11-2. HP-IB Bus Structure

The data bus consists of 8 bi-directional lines that are used to transfer data from one device to

another . Programming commands and data transmitted on these lines are typically encoded

in ASCII, although binary encoding is often used to speed up the transfer of large arrays .

Both ASCII- and binary-data formats are available to the analyzer . In addition, every byte

transferred over HP-IB undergoes a handshake to insure valid data.

A three-line handshake scheme coordinates the transfer of data between talkers and listeners .

T o insure data integrity in multiple-listener transfers , this technique forces data transfers

to occur at the transfer rate of the slowest device connected to the interface . With most

computing controllers and instruments , the handshake is performed automatically , making it

transparent to the programmer .

The data bus also has ve control lines . The controller uses these lines to address devices and

to send bus commands .

IFC (Interface Clear) This line is used exclusively by the system controller . When

this line is true (low), all devices (whether addressed or not)

unaddress and revert to an idle state .

11-16 Compatible P eripherals

Show 4 more labels / short text on this page
HP-IB Bus Structure · Data Bus · Handshake Lines · Control Lines
Page 625 of the Agilent Technologies 8753D user manual — LOCAL
Page 625 · click to enlarge
Page 625

A TN (Attention) The active controller uses this line to de ne whether

the information on the data bus is command-oriented or

data-oriented. When this line is true (low), the bus is in the

command mode , and the data lines carry bus commands . When

this line is false (high), the bus is in the data mode , and the

data lines carry device-dependent instructions or data.

SRQ (Service Request) This line is set true (low) when a device requests service

and the active controller services the requesting device . The

network analyzer can be enabled to pull the SRQ line for a

variety of reasons such as requesting control of the interface ,

for the purposes of printing, plotting, or accessing a disk.

REN (Remote Enable) This line is used exclusively by the system controller . When this

line is set true (low), the bus is in the remote mode , and devices

are addressed by the controller to either listen or talk. When

the bus is in remote mode and a device is addressed, it receives

instructions from the system controller via HP-IB rather than

from its front panel (pressing

returns the device to

LOCAL

front-panel operation). When this line is set false (high), the

bus and all of the connected devices return to local operation.

EOI (End or Identify) This line is used by a talker to indicate the last data byte in a

multiple-byte transmission, or by an active controller to initiate

a parallel-poll sequence . The analyzer recognizes the EOI line

as a terminator , and it pulls the EOI line with the last byte of a

message output (data, markers , plots , prints , error messages).

The analyzer does not respond to parallel poll.

Number of Interconnected Devices: 15 maximum.

Interconnection P ath Maximum Cable Length: 20 meters maximum or 2 meters per device

(whichever is less).

Message Transfer SchemeByte serial, bit parallel a synchronous data

transfer using a 3-line handshake system.

Data RateMaximum of 1 megabyte-per-second over the

speci ed distances with tri-state drivers . A ctual

data rate depends on the transfer rate of the

slowest device connected to the bus .

A ddress CapabilityPrimary addresses: 31 talk, 31 listen. A

maximum of 1 talker and 14 listeners can be

connected to the interface at given time .

Multiple-Controller CapabilityIn systems with more than one controller (like

this instrument), only one controller can be

active at any given time . The active controller

can pass control to another controller , but only

the system controller can assume unconditional

control. Only one system controller is allowed.

Compatible P eripherals 11-17

HP-IB Requirements

Page 626 of the Agilent Technologies 8753D user manual — Chapter 11 Compatible Peripherals
Page 626 · click to enlarge
Page 626

On the network analyzer's rear panel, next to the HP-IB connector , there is a list of HP-IB

device subsets as de ned by the IEEE 488.2 standard. The analyzer has the following

capabilities:

SH1 Full-source handshake .

AH1 Full-acceptor handshake .

T6 Basic talker , answers serial poll, unaddresses if MLA is issued. No talk-only mode .

L4 Basic listener , unaddresses if MT A is issued. No listen-only mode .

SR1 Complete service request (SRQ) capabilities .

RL1 Complete remote/local capability including local lockout.

PP0 Does not respond to parallel poll.

DC1 Complete device clear .

DT1 Responds to a Group Execute Trigger (GET) in the hold-trigger mode .

C1,C2,C3 System controller capabilities in system-controller mode .

C10 P ass control capabilities in pass-control mode .

E2 Tri-state drivers .

LE0 No extended listener capabilities .

TE0 No extended talker capabilities .

These codes are completely explained in the IEEE Std 488 documents , published by the

Institute of Electrical and Electronic Engineers , Inc ., 345 East 47th Street, New Y ork,

New Y ork 11017.

11-18 Compatible P eripherals

HP-IB Operational Capabilities

Page 627 of the Agilent Technologies 8753D user manual — Chapter 11 Compatible Peripherals
Page 627 · click to enlarge
Page 627

When the analyzer is connected to other instruments over the HP-IB , the HP-IB status

indicators illuminate to display the current status of the analyzer . The HP-IB status indicators

are located in the instrument-state function block on the front panel of the network analyzer .

R = Remote Operation

L = Listen mode

T = T alk mode

S = Service request (SRQ) asserted by the analyzer

The analyzer uses a single-bus architecture . The single bus allows both the analyzer and the

host controller to have complete access to the peripherals in the system.

Three di erent controller modes are possible in and HP-IB system:

Agilent Technologies 8753D — Chapter 11 Compatible Peripherals diagram (page 627)
Fig. 260 · Chapter 11 Compatible Peripherals

Figure 11-3. Analyzer Single Bus Concept

Compatible P eripherals 11-19

Show 5 more labels / short text on this page
HP-IB Status Indicators · Bus Device Modes · system-controller mode · talker/listener mode · pass-control mode
Page 628 of the Agilent Technologies 8753D user manual — LOCAL
Page 628 · click to enlarge
Page 628

This mode allows the analyzer to control peripherals directly in a stand-alone environment

(without an external controller). This mode can only be selected manually from the

analyzer's front panel. It can only be used if no active computer or instrument controller is

connected to the system via HP-IB . If an attempt is made to set the network analyzer to the

system-controller mode when another controller is connected to the interface , the following

message is displayed on the analyzer's display screen:

"ANOTHER SYSTEM CONTROLLER ON HP-IB BUS"

The analyzer must be set to the system-controller mode in order to access peripherals from the

front panel. In this mode , the analyzer can directly control peripherals (plotters , printers , disk

drives , power meters , etc .) and the analyzer may plot, print, store on disk or perform power

meter functions .

Do not attempt to use this mode for programming. HP recommends using an

external instrument controller when programming. See the following section,

\T alker/Listener Mode ."

This is the mode that is normally used for remote programming of the analyzer . In

talker/listener mode , the analyzer and all peripheral devices are controlled from an external

instrument controller . The controller can command the analyzer to talk and other devices

to listen. The analyzer and peripheral devices cannot talk directly to each other unless the

computer sets up a data path between them. This mode allows the analyzer to act as either

a talker or a listener , as required by the controlling computer for the particular operation in

progress .

This mode allows the computer to control the analyzer via HP-IB (as with the talker/listener

mode), but also allows the analyzer to take control of the interface in order to plot, print, or

access a disk. During an analyzer-controlled peripheral operation, the host computer is free to

perform other internal tasks (i.e . data or display manipulation) while the analyzer is controlling

the bus . After the analyzer-controlled task is completed, the analyzer returns control to the

system controller .

Setting HP-IB A ddresses

In systems interfaced using HP-IB , each instrument on the bus is identi ed by an HP-IB

address . This address code must be di erent for each instrument on the bus . These addresses

are stored in short-term, non-volatile memory and are not a ected when you press

PRESET

cycle the power . The analyzer occupies two HP-IB addresses: the instrument itself and the

display . The display address is derived from the instrument address by complementing the

instrument's least-signi cant bit. Hence , if the instrument is at an even address , the display

occupies the next higher address . If the instrument is at an odd address , the display occupies

the next lower address .

The analyzer addresses are set by pressing

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SET ADDRESSES . In system-controller mode ,

LOCAL

the addresses must be set for the plotter , printer , disk drive , and power meter .

The default address for the analyzer is device 16, and the display address is device 17.

11-20 Compatible P eripherals

Show 5 more labels / short text on this page
System-Controller Mode · T alker/Listener Mode · P ass-Control Mode · or ·
Page 629 of the Agilent Technologies 8753D user manual — Chapter 11 Compatible Peripherals
Page 629 · click to enlarge
Page 629

There is also an address for the system controller . This address refers to the

controller when the network analyzer is being used in pass-control mode . This

is the address that control is passed back to when the analyzer-controlled

operation is complete .

The analyzer HP-IB commands are derived from their front-panel key titles (where possible),

according to this naming convention:

Simple commands are the rst four letters of the function they control, as in POWE, the

command name for power . If the function label contains two words , the rst three mnemonic

letters are the rst three letters of the rst word, and the fourth mnemonic letter is the rst

letter of the second word. F or example , ELED is derived from electrical delay .

If there are many commands grouped together in a category , as in markers or plotting pen

numbers , the command is increased to 8 letters . The rst 4 letters are the category label

derived using rule 1. The last 4 letters are the function speci er , again derived using rule 1. As

an example , category pen numbers are represented by the command PENN, which is used in

combination with several functions such as PENND A T A, PENNMEMO .

The code naming guidelines , listed in T able 11-1, are used in order to:

make commands more meaningful and easier to remember

maintain compatibility with other products (including the HP 8510)

There are times when these guidelines are not followed due to technical

considerations .

Compatible P eripherals 11-21

Analyzer Command Syntax · Code Naming Convention

Page 630 of the Agilent Technologies 8753D user manual — CALKN50
Page 630 · click to enlarge
Page 630

T able 11-1. Code Naming Convention

Convention K ey Title F or HP-IB Code Use Example

One W ord P ower First F our Letters POWE

Two W ords Electrical Delay First Three Letters of First W ord, First Letter

ELED

of Second W ord

Two W ords in a Group Marker !Center F our Letters of Both MARKCENT

Gate !Span GA TESP AN

Three W ords Cal Kit N 50 First Three Letters of First W ord, First Letter

CALKN50

of Second W ord, First F our Letters of Third

P en Num Data PENND A T A

Some codes require appendages (ON, OFF , 1, 2, etc .). Codes that do not have a front-panel

equivalent are HP-IB only commands . They use a similar convention based on the common

name of the function.

The analyzer accepts the following ASCII characters:

Both upper- and lower-case letters are acceptable . Carriage returns , leading zeros , spaces , and

unnecessary terminators are ignored, except for those within a command or appendage . If the

analyzer does not recognize a character as appropriate , it generates a syntax error message and

recovers at the next terminator .

11-22 Compatible P eripherals

Show 12 more labels / short text on this page
Start ST AR · Search Right SEAR · W ord · V alid Characters · letters · numbers · decimal points · +/ · semicolons (;) · quotation marks (") · carriage returns (CR) · linefeeds (LF)
Page 631 of the Agilent Technologies 8753D user manual — Chapter 11 Compatible Peripherals
Page 631 · click to enlarge
Page 631

The analyzer can input and output data in basic units such as Hz, dB , seconds , ohms , etc .

S Seconds HZ Hertz

V V olts DB dB or dBm

Input data is assumed to be in basic units unless one of the following units expressions quali es

the data input (upper and lower case are equivalent):

MS Milliseconds kHz Kilohertz

US Microseconds MHz Megahertz

NS Nanoseconds GHz Gigahertz

PS Picoseconds FS F emtoseconds

An HP-IB diagnostic feature (debug mode) is available in the HP-IB menu. A ctivating the debug

mode causes the analyzer to scroll incoming HP-IB commands across the display . Nonprintable

characters are represented with a . Any time the analyzer receives a syntax error , the

commands halt, and a pointer indicates the misunderstood character . See the HP 8753D

Network Analyzer Programmer's Guide for information on correct programming syntax.

Compatible P eripherals 11-23

Units · HP-IB Debug Mode

Page 632 of the Agilent Technologies 8753D user manual — Chapter 11 Compatible Peripherals
Page 632 · click to enlarge
Page 632
Page 632 of the Agilent Technologies 8753D user manual — Chapter 11 Compatible Peripherals
Page 633 of the Agilent Technologies 8753D user manual — Chapter 12 Preset State and Memory Allocation
Page 633 · click to enlarge
Page 633

Preset State and Memory Allocation

The analyzer is capable of saving complete instrument states for later retrieval. It can store

these instrument states into the internal memory , to the internal disk, or to an external disk.

This chapter describes these capabilities in the following sections:

instrument state de nition

internal and external data storage

description of analyzer state after preset

Where to Look for More Information

A dditional information about many of the topics discussed in this chapter is located in the

following areas:

Chapter 2, \Making Measurements ," contains step-by-step procedures for making

measurements or using particular functions .

Chapter 4, \Printing, Plotting, and Saving Measurement Results ," contains instructions

for saving to disk or the analyzer internal memory , and printing and plotting displayed

measurements .

Types of Memory and Data Storage

The analyzer utilizes two types of memory and can also utilize the internal disk drive or be

connected to an external disk drive:

This is dynamic read/write memory , of approximately 2 Mbytes , that contains all of the

parameters that make up the current instrument state . An instrument state consists of all the

stimulus and response parameters that set up the analyzer to make a speci c measurement.

Some data that you may think is part of the instrument state (such as calibration data and

memory traces) are actually stored in non-volatile memory . See \Non-V olatile Memory" to read

more about the di erences .

V olatile memory is cleared upon a power cycle of the instrument and, except as noted, upon

instrument preset.

Preset State and Memory Allocation 12-1

memory allocation · V olatile Memory

Page 634 of the Agilent Technologies 8753D user manual — PRESET
Page 634 · click to enlarge
Page 634

This is CMOS read/write memory that is protected by a battery to provide storage of data when

line power to the instrument is turned o . With this battery protection, data can be retained in

memory for 250 days at 70

C and for 10 years at 25

C (characteristically).

Non-volatile memory consists of a block of user-allocated memory and a block of xed memory .

The user-allocated memory is available for you to save the following data:

power meter calibration data

user calibration kit de nitions

Even though calibration data is stored in non-volatile memory , if the associated

instrument state is not saved, you will not be able to retrieve the calibration

data after a power cycle .

The xed memory is used to store the following data (you cannot change where this data is

stored and it does not a ect your memory availability for storing user-allocated data):

copy con guration (printer and plotter type , port, baud rate , handshake)

power meter type (HP 436/438)

power sensor calibration factors and loss tables

user-de ned calibration kits

The maximum number of instrument states , calibrations , and memory traces that can reside

in non-volatile memory at any one time is limited to 31 instrument states , 128 calibrations (4

per instrument state , including the present instrument state), and 64 memory traces (2 per

instrument state , including the present instrument state).

In addition, the number of instrument states and associated calibrations and memory traces are

limited by the available memory . T o display the amount of unused memory on the analyzer ,

. (Be sure you have selected

N

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SA VE/RECALL

INTERNAL MEMORY as your disk type .) In the

upper right-hand portion of the display , the value displayed as Bytes free: is the unused

non-volatile memory . When you save to the internal memory , you will see the number of bytes

free decrease . When you delete les , the number of bytes free increases . There is a maximum

of 512 kBytes available .

If you have deleted registers since the last time the instrument was preset, the bytes available

for you to use may be less than the actual \bytes free" that is displayed. Deleting registers

to increase the available memory will work in cases where the registers being deleted and

the registers needing to be added are of the same standard size (such as instrument states not

having calibrations associated with them). In certain other cases , however , you may have

after deleting registers so that the \bytes free" value equals the available

PRESET

memory value . During a preset, the analyzer runs a memory packer that de-fragments the free

memory into one contiguous block.

12-2 Preset State and Memory Allocation

Show 17 more labels / short text on this page
Non-V olatile Memory · · · instrument states · measurement calibration data · memory traces · user preset · HP-IB addresses · display colors · sequence titles · sixth sequence · system Z0 · factory preset · HP-IB con guration · display intensity default · press · to press
Page 635 of the Agilent Technologies 8753D user manual — PRESET
Page 635 · click to enlarge
Page 635

T able 12-1 shows the memory requirements of calibration arrays and memory trace arrays

to help you approximate memory requirements . F or example , add the following memory

requirements:

a full 2-port calibration with 801 points (58 k)

the memory trace array (4.9 k)

the instrument state (3 k)

The total memory requirement is 65.9 kbytes . There is sucient memory to store 7 calibrations

of this type . However , the same calibration performed with 1601 points and 2 channels

uncoupled would require 252 k bytes:

a full 2-port calibration with 1601 points , two channels , uncoupled (230 k)

the memory trace array (19 k)

the instrument state (3 k)

Only 2 of these calibrations could reside in memory before the available memory would be

depleted.

T able 12-1.

Memory Requirements of Calibration and Memory Trace Arrays

V ariable Data Length (Bytes) Approximate T otals (Bytes)

401 pts 801 pts 1601 pts

1 chan 1 chan 2 chans

Response N 6 + 52 2.5 k 5 k 10 k 19 k

Response and isolation N 6 2 + 52 5 k 10 k 19 k 38 k

1-P ort N 6 3 + 52 7 k 14 k 29 k 58 k

2-P ort N 6 12 + 52 29 k 58 k 115 k 230 k

Interpolated cal Same as above in addition to regular cal

P ower Meter Cal

(N

2 number of channels

) +208 1 k 1.8 k 3.4 k 6.6 k

N 6 + 52 2.5 k 4.9 k 9.7 k 19 k

3 k 3 k 3 k 3 k

N = number of points

This variable is allocated once per active channel.

The number of points that was set at the time the cal was turned on.

If the channels are coupled, this number is always 1. If the channels are uncoupled, this number

refers to the number of channels that have power meter cal on.

This value may change with di erent rmware revisions .

The analyzer attempts to allocate memory at the start of a calibration. If insucient memory

is available , an error message is displayed. It is possible that the CMOS memory might be

fragmented due to the sequence of saving and deleting states of various sizes . So another

alternative would be to store the current state to disk and then press

. The analyzer

PRESET

runs a memory packer which might regain some previously inaccessible memory . If memory is

still inadequate , delete an instrument state and restart the calibration.

Preset State and Memory Allocation 12-3

Show 13 more labels / short text on this page
Calibration Arrays · * · y · z · Measurement Data · Memory trace array · * · Instrument State · # · * · y · z · #
Page 636 of the Agilent Technologies 8753D user manual — EXTERNAL DISK
Page 636 · click to enlarge
Page 636

Storing Data to Disk

Y ou can use the internal disk drive or connect an external disk drive for storage of instrument

states , calibration data, measurement data, and plot les . (Refer to Chapter 4, \Printing,

Plotting, and Saving Measurement Results", for more information on saving measurement data

and plot les .)

The analyzer displays one le name per stored instrument state when you list the disk

directory . In reality , several les are actually stored to the disk when you store the instrument

state . Thus , when the disk directory is accessed from a remote system controller , the directory

will show several les associated with a particular saved state . The maximum number of les

that you can store on a disk depends on the directory size . Y ou can de ne the directory size

when you format a disk. See T able 12-3 for the default directory size for oppy disks and hard

disks .

The maximum number of instrument states and calibrations that can reside on a disk is

limited by the available disk space . T o see the available disk space displayed on the analyzer ,

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SA VE/RECALL

. (Be sure you have selected either

EXTERNAL DISK

depending on your disk type .) In the upper righthand portion of the display , the value

displayed as Bytes free: is the available disk space . If your disk is formatted in LIF , this

value is the largest contiguous block of disk space . Since the analyzer is reporting the largest

contiguous block of disk space , you may or may not see the bytes free number change when

you delete les . If your disk is formatted in DOS, the number reported as bytes free is the total

available disk space . That number is updated whenever you save to or delete les from the

disk.

A disk le created by the analyzer appends a sux to the le name . (This is on the analyzer's

directory and is not visible .) The sux consists of one or two characters: the rst character is

the le type and the second is a data index. (Refer to T able 12-2 for the de nitions of each

sux character .)

T able 12-2. Sux Character De nitions

Char 1 De nition Char 2 De nition

G Graphics 1 Display graphics

D Error corrected data 1 Channel 1

R Raw data 1 to 4 Channel 1, raw arrays 1 to 4

5 to 8 Channel 2, raw arrays 5 to 8

F F ormatted data 1 Channel 1

C Cal K Cal kit

1 Cal data, channel 1 0 Stimulus state

1 to 9 Coecients 1 to 9

2 Cal data, channel 2 0 to C same as channel 1

M Memory trace data 1 Channel 1

If correction is on at the time of an external store , the calibration set is stored to disk. (Note

that inactive calibrations are not stored to disk.) When an instrument state is loaded into the

12-4 Preset State and Memory Allocation

Show 10 more labels / short text on this page
press · INTERNAL DISK or · I Instrument state · 0 Graphics index · 2 Channel 2 · 2 Channel 2 · A Coecient 10 · B Coecient 11 · C Coecient 12 · 2 Channel 2
Page 637 of the Agilent Technologies 8753D user manual — Chapter 12 Preset State and Memory Allocation
Page 637 · click to enlarge
Page 637

analyzer from disk, the stimulus and response parameters are restored rst. If correction is on

for the loaded state , the analyzer will load a calibration set from disk that carries the same title

as the one stored for the instrument state .

If you are concerned about conserving memory , either internal memory or external disk space ,

some of the most memory-intensive operations include:

saving data arrays and graphics with the instrument state

Using Saved Calibration Sets

When you are saving to internal memory (CMOS, non-volatile memory), calibration sets are

linked to the instrument state and measurement parameter for which the calibration was

done . Therefore a saved calibration can be used for multiple instrument states as long as the

measurement parameter , frequency range , and number of points are the same . A full 2-port

calibration is valid for any S-parameter measurement with the same frequency range and

number of points . When an instrument state is deleted from memory , the associated calibration

set is also deleted if it is unused by any other state .

The following hints will help you avoid potential problems:

If a measurement is saved with calibration and interpolated calibration on, it will be restored

with interpolated calibration on.

A calibration stored from one instrument and recalled by a di erent one will be invalid. T o

ensure maximum accuracy , always recalibrate in these circumstances .

No record is kept in memory of the temperature when a calibration set was stored.

Instrument characteristics change as a function of temperature , and a calibration stored at

one temperature may be inaccurate if recalled and used at a di erent temperature . See

Chapter 7, \Speci cations and Measurement Uncertainties ," for allowable temperature ranges

for individual speci cations .

The HP 8753D can read disk les created by the HP 8753B/C and the HP 8753B/C can read

les created by the HP 8753D . A disk le translator is available to make HP 8753A disk les

compatible with HP 8753B les . These les can then be read by the HP 8753D . Contact your

local Hewlett-P ackard sales and service oce for a copy of this disk le translator .

Preset State and Memory Allocation 12-5

Show 5 more labels / short text on this page
Conserving Memory · two-port error correction · interpolated error correction · 1601 measurement points · using time domain
Page 638 of the Agilent Technologies 8753D user manual — PRESET
Page 638 · click to enlarge
Page 638

key is pressed, the analyzer reverts to a known state called the factory

PRESET

preset state . This state is de ned in T able 12-3. There are subtle di erences between the

preset state and the power-up state . These di erences are documented in T able 12-4 . If power

to non-volatile memory is lost, the analyzer will have certain parameters set to default settings .

T able 12-5 shows the a ected parameters .

When line power is cycled, or the

key pressed, the analyzer performs a self-test

PRESET

routine . Upon successful completion of that routine , the instrument state is set to the

conditions shown in T able 12-3 . The same conditions are true following a \PRES;" or \RST;"

command over HP-IB , although the self-test routines are not executed.

Y ou also can con gure an instrument state and de ne it as your user preset state:

1. Set the instrument state to your desired preset conditions .
2. Save the state (save/recall menu).
3. Rename that register to \UPRESET".

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRESET

PRESET:USER .

key is now toggled to the

NNNNNNNNNNNNNN

USER selection and your de ned instrument state will

PRESET

be recalled each time you press

and when you turn power on. Y ou can toggle back to

PRESET

N

NNNNNNNNNNNNNNNNNNNNNN

the factory preset instrument state by pressing

PRESET

FACTORY .

When you send a preset over HP-IB , you will always get the factory preset.

Y ou can, however , activate the user-de ned preset over HP-IB by recalling the

register in which it is stored.

12-6 Preset State and Memory Allocation

Show 6 more labels / short text on this page
Preset State · When the · Press · · The · and selecting
Page 639 of the Agilent Technologies 8753D user manual — PRESET
Page 639 · click to enlarge
Page 639

T able 12-3. Preset Conditions (1 of 5)

Preset Conditions Preset V alue

Preset Conditions Preset V alue

Analyzer Mode Network Analyzer Mode

Frequency O set O

Edit Mode Start/Stop , Number of

O set V alue 0

P arameter Channel 1S11;
Channel 2S21

Sweep Type Linear Frequency

F ormat Log Magnitude (all inputs)

Color Selections Same as before

PRESET

Sweep Time 100 ms , A uto Mode

Start Frequency 30 kHz

A ctive Channel Channel 1

Frequency Span (1.) 2999.97 MHz

Frequency Span (Opt. 006) 5999.97 MHz

Intensity If set to 15%,

PRESET

Time Span 100 ms

has no e ect. If set to

CW Frequency 1000 MHz

PRESET

Source P ower 0 dBm

P ower Slope 0 dB/GHz; O

to 15%.

Start P ower 15.0 dBm

BeeperDone On

P ower Span 25 dB

BeeperW arning O

Coupled P ower On

D2/D1 to D2 O

Source P ower On

Title Channel 1 = [hp]

Channel 2 = Empty

Coupled P ort P ower On

IF Bandwidth 3000 Hz

P ower Range A uto; Range 0

IF A veraging F actor 16; O

Number of P oints 201

Smoothing Aperture 1% SP AN; O

Phase O set 0 Degrees

Electrical Delay 0 ns

Preset State and Memory Allocation 12-7

Show 24 more labels / short text on this page
Analyzer Mode · Frequency List · Frequency List Empty · Operation · P oints · Harmonic Operation O · Response Conditions · Stimulus Conditions · Conversion O · Display Mode Start/Stop · Trigger Type Continuous · Display Data · External Trigger O · · Dual Channel O · Frequency Blank Disabled · Split Display On · Start Time 0 · · < 15% · · increases intensity · Coupled Channels On · Scale/Division 10 dB/Division
Page 640 of the Agilent Technologies 8753D user manual — Chapter 12 Preset State and Memory Allocation
Page 640 · click to enlarge
Page 640

T able 12-3. Preset Conditions (2 of 5)

Preset Conditions Preset V alue

Preset Conditions Preset V alue

P olar Marker Lin Mkr

Smith Marker R+jX Mkr

Calibration Kit 7 mm

System Z0 50 Ohms

V elocity F actor 1

P ort 1 0 s

Limit T esting O

P ort 2 0 s

Input A 0 s

Edit Mode Upper/Lower Limits

Input B 0 s

Stimulus O set 0 Hz

Chop A and B On

Amplitude O set 0 dB

P ower Meter Calibration O

Limit Type Sloping Line

Number of Readings 1

Beep F ail O

P ower Loss Correction O

Start Transform 20 nanoseconds

Transform Span 40 nanoseconds

Markers 1, 2, 3, 4, 5 1 GHz; All Markers O

Last A ctive Marker 1

Gate Start 10 nanoseconds

Gate Span 20 nanoseconds

Delta Marker Mode O

Marker T arget V alue 3 dB

Marker Width V alue 3 dB; O

HP-IB A ddresses Last A ctive State

HP-IB Mode Last A ctive State

Marker Stimulus O set 0 Hz

F ocus Last A ctive State

Marker V alue O set 0 dB

Clock Time Stamp On

Marker A ux O set (Phase) 0 Degrees

PresetF actory/User Last Selected State

12-8 Preset State and Memory Allocation

Show 27 more labels / short text on this page
Calibration · Marker Statistics O · Correction O · Calibration Type None · Limit Lines · Extensions O · Limit Lines O · Limit List Empty · Sensor A/B A · Time Domain · Interpolated Error O · Transform O · Correction · Transform Type Bandpass · Markers (coupled) · Gating O · Gate Shape Normal · Reference Marker None · Marker Mode Continuous · Display Markers On · Demodulation O · Window Normal · Coupling On · Use Memory O · Marker Search O · System P arameters · Marker Tracking O
Page 641 of the Agilent Technologies 8753D user manual — Chapter 12 Preset State and Memory Allocation
Page 641 · click to enlarge
Page 641

T able 12-3. Preset Conditions (3 of 5)

Preset Conditions Preset V alue

Preset Conditions Preset V alue

P arallel P ort Last A ctive State

Spur A voidance On

Plotter Type Last A ctive State

A ux Input Resolution Low

Plotter P ort Last A ctive State

Analog Bus Node 11 (A ux Input)

Plotter Baud Rate Last A ctive State

Plotter Handshake Last A ctive State

HP-IB A ddress Last A ctive State

Printer Type Last A ctive State

Printer P ort Last A ctive State

Printer Baud Rate Last A ctive State

Plot T ext On

Printer Handshake Last A ctive State

Printer HP-IB A ddress Last A ctive State

Plot Quadrant Full P age

Disk Save Con guration

Plot Speed F ast

P en Number:

Raw Data Array O

F ormatted Data Array O

Select Disk Internal Memory

Ch1 T ext 7

Disk F ormat LIF

Ch2 T ext 7

Line Type:

Source Phase Lock Loop On

1 The directory size is calculated as 0.013% of the oppy disk size (which is 256) or 0.005% of the hard disk size .

2 Pressing preset turns o sequencing modify (edit) mode and stops any running sequence .

Preset State and Memory Allocation 12-9

Show 32 more labels / short text on this page
Copy Con guration · Sampler Correction On · Plot · Plot Data On · Plot Memory On · Plot Graticule On · Plot Marker On · A utofeed On · Scale Plot Full · (De ne Store) · Data Array O · Ch1 Data 2 · Ch2 Data 3 · Graphics O · Ch1 Memory 5 · Data Only O · Ch2 Memory 6 · Directory Size Default · Ch1 Graticule 1 · Save Using Binary · Ch2 Graticule 1 · Ch1 Marker 7 · Sequencing · Ch2 Marker 7 · Loop Counter 0 · TTL OUT High · Ch1 Data 7 · Ch2 Data 7 · Service Modes · Ch1 Memory 7 · HP-IB Diagnostic O · Ch2 Memory 7
Page 642 of the Agilent Technologies 8753D user manual — SWR 1.00 0.0 1.0
Page 642 · click to enlarge
Page 642

T able 12-3. Preset Conditions (4 of 5)

Preset Conditions Preset V alue

Preset Conditions Preset V alue

Printer Mode Last A ctive State

A uto-F eed On

T able 12-3. Preset Conditions (5 of 5)

F ormat T able Scale

P osition V alue

Log Magnitude (dB) 10.0 5.0 0.0

Phase (degree) 90.0 5.0 0.0

Group Delay (ns) 10.0 5.0 0.0

Smith Chart 1.00 { 1.0

P olar 1.00 { 1.0

Linear Magnitude 0.1 0.0 0.0

Real 0.2 5.0 0.0

Imaginary 0.2 5.0 0.0

SWR 1.00 0.0 1.0

T able 12-4. P ower-on Conditions (versus Preset)

HP-IB MODE T alker/listener .

SA VE REGISTERS P ower meter calibration data and calibration data not associated with

an instrument state are cleared.

COLOR DISPLA Y Default color values .

INTENSITY F actory stored values . The factory values can be changed by running

the appropriate service routine . Refer to the \A djustments and

Correction Constants" chapter in the HP 8753D Service Guide.

SEQUENCES Sequence 1 through 5 are erased.

DISK DIRECTORY Cleared.

12-10 Preset State and Memory Allocation

Show 10 more labels / short text on this page
Print · CH2 Data Blue · CH2 Mem Red · Graticule Cyan · Printer Colors · W arning Black · CH1 Data Magenta · T ext Black · CH1 Mem Green · Reference
Page 643 of the Agilent Technologies 8753D user manual — EXTERNAL REGISTER TITLES
Page 643 · click to enlarge
Page 643

T able 12-5. Results of P ower Loss to Non-V olatile Memory

HP-IB ADDRESSES are set to the following defaults:

HP 8753D : 16

USER DISPLA Y : 17

PLOTTER : 5

PRINTER : 1

POWER METER : 13

DISK : 0

DISK UNIT NUMBER : 0

DISK V OLUME NUMBER : 0

POWER METER TYPE is set to HP 438A/437

INTERNAL REGISTER TITLES

are set to defaultsREG1 through REG32

EXTERNAL REGISTER TITLES

(store les) are set to defaults: FILE1 through FILE 5

PRINT TYPE is set to default: MONOCHROME

PRINTING/PLOTTING SETUPS are set to the following defaults:

P ARALLEL PORT : COPY

PLOTTER TYPE : PLOTTER

PLOTTER PORT : SERIAL

PLOTTER B A UD RA TE : 9600

PLOTTER HANDSHAKE : Xon-Xo

PRINTER TYPE : DESKJET

PRINTER PORT : P ARALLEL

PRINTER B A UD RA TE : 19200

PRINTER HANDSHAKE : Xon-Xo

1 Only applies to HP-IB operation.

Preset State and Memory Allocation 12-11

Page 644 of the Agilent Technologies 8753D user manual — Chapter 12 Preset State and Memory Allocation
Page 644 · click to enlarge
Page 644
Page 644 of the Agilent Technologies 8753D user manual — Chapter 12 Preset State and Memory Allocation
Page 645 of the Agilent Technologies 8753D user manual — A
Page 645 · click to enlarge
Page 645

A

The CITI le Data F ormat and K eyword Reference

This appendix contains the following information:

The CITI le Data F ormat

De nition Of CITI le T erms

The CITI le Keyword Reference .

The CITI le Data F ormat

CITI le is a standardized data format, used for exchanging data between di erent computers

and instruments . CITI le is an abbreviation for \Common Instrumentation Transfer and

Interchange le". This standard has been a group e ort between instrument designers and

designers of computer-aided design programs . As much as possible , CITI le meets current needs

for data transfer , and it was designed to be expandable so it can meet future needs .

CITI le de nes how the data inside an ASCII package is formatted. Since it is not tied to any

particular disk or transfer format, it can be used with any operating system (B ASIC, DOS,

UNIX, etc .), with any disk format (LIF , DOS, 1, etc .), or with any transfer mechanism (disk,

LAN, GPIB , etc .).

By careful implementation of the standard, instruments and software packages using CITI le

are able to load and work with data created on another instrument or computer . It is possible ,

for example , for a network analyzer to directly load and display data measured on a scalar

analyzer , or for a software package running on a computer to read data measured on the

network analyzer .

There are two main types of data formats: binary and ASCII. CITI le uses the ASCII text

format. While this format does take up more bytes of space than a binary format, ASCII data

is a transportable , standard type of format which is supported by all operating systems . In

addition, the ASCII format is accepted by most text editors . This allows les to be created,

examined, and edited easily , making CITI le easier to test and debug.

File and Operating System F ormats

CITI le was designed to be independent of the data storage mechanism, and therefore may

be implemented for any le system. However transfer between le systems may sometimes

be necessary . Any commercially available software that has the ability to transfer ASCII les

between systems may be used to transfer CITI le data.

The CITI le Data Format and K eyword Reference A-1

Show 4 more labels / short text on this page
Description and Overview · CITI le Examples · Description and Overview · Data F ormats
Page 646 of the Agilent Technologies 8753D user manual — -3.54545E-2,-1.38601E-3
Page 646 · click to enlarge
Page 646

De nition of CITI le T erms

This section will de ne the following terms:

A CITI le P ackage

A typical package is divided into two parts: The rst part, the header , is made up of keywords

and setup information. The second part, the data, usually consists of one or more arrays of

data. Example 1 shows the basic structure of a CITI le package:

Example 1, A CITI le P ackage

The \header" part CITIFILE A.01.00

NAME MEMORY

V AR FREQ MA G 3

D A T A S RI

The \data" part BEGIN

-3.54545E-2,-1.38601E-3

0.23491E-3,-1.39883E-3

2.00382E-3,-1.40022E-3

END

When stored in a disk le there may be more than one CITI le package . With the 8510 network

analyzer , for example , storing a \memory all" will save all eight of the memories held in the

instrument. This results in a single le which contains eight CITI le packages .

The CITI le Header

The header section contains information about the data that will follow . It may also include

information about the setup of the instrument that measured the data. The CITI le header

shown in Example 1 has just the bare minimum of information necessary; no instrument setup

information was included.

An Array of Data

An array is numeric data that is arranged with one data element per line . A CITI le package

may contain more than one array of data. Arrays of data start after the BEGIN keyword, and

the END keyword will follow the last data element in an array .

A CITI le package does not necessarily need to include data arrays; for instance , CITI le could

be used to store the current state of an instrument. In that case the keywords V AR, D A T A,

BEGIN, and END would not be required.

A-2 The CITI le Data Format and K eyword Reference

Show 4 more labels / short text on this page
package · header · data array · keyword
Page 647 of the Agilent Technologies 8753D user manual — A - The CITIfile Data Format and Keyword Reference
Page 647 · click to enlarge
Page 647

CITI le K eyword

Keywords are always the rst word on a new line . They are always one continuous word

without embedded spaces .

A listing of all the keywords used in the latest A.01.01 version of CITI le is shown in \The

CITIFile Keyword Reference ."

When reading a CITI le , unrecognized keywords should be ignored. This allows new keywords

to be added, without a ecting an older program or instrument that might not use the new

keywords . The older instrument or program can still use the rest of the data in the CITI le as it

did before . Ignoring unknown keywords allows backwards compatibility to be maintained.

The CITI le Data Format and K eyword Reference A-3

Page 648 of the Agilent Technologies 8753D user manual — CITIFILE A.01.00
Page 648 · click to enlarge
Page 648

Example 2, An 8510 Display Memory File

Example 2 shows a simple le that contains no frequency information. Some instruments do

not keep frequency information for display memory data, so this information is not included in

the CITI le package .

in this le . This convention allows the designer to de ne keywords that are particular to his or

her particular implementation.

CITIFILE A.01.00

#NA VERSION HP8510B.05.00

NAME MEMORY

#NA REGISTER 1

VAR FREQ MAG 5

DATA S RI

BEGIN

-1.31189E-3,-1.47980E-3

-3.67867E-3,-0.67782E-3

-3.43990E-3,0.58746E-3

-2.70664E-4,-9.76175E-4

0.65892E-4,-9.61571E-4

END

Example 3, 8510 Data le

Example 3 shows a CITI le package created from the data register of an 8510 Network

Analyzer . In this case 10 points of real and imaginary data was stored, and frequency

information was recorded in a segment list table .

CITIFILE A.01.00

#NA VERSION HP8510B.05.00

NAME DATA

#NA REGISTER 1

VAR FREQ MAG 10

DATA S[1,1] RI

SEG_LIST_BEGIN

SEG 1000000000 4000000000 10

SEG_LIST_END

BEGIN

0.86303E-1,-8.98651E-1

8.97491E-1,3.06915E-1

-4.96887E-1,7.87323E-1

-5.65338E-1,-7.05291E-1

8.94287E-1,-4.25537E-1

1.77551E-1,8.96606E-1

-9.35028E-1,-1.10504E-1

3.69079E-1,-9.13787E-1

7.80120E-1,5.37841E-1

-7.78350E-1,5.72082E-1

END

A-4 The CITI le Data Format and K eyword Reference

CITI le Examples · Example · Example

Page 649 of the Agilent Technologies 8753D user manual — CITIFILE A.01.00
Page 649 · click to enlarge
Page 649

Example 4, 8510 3-T erm Frequency List Cal Set File

Example 4 shows how CITI le may be used to store instrument setup information. In the case

of an 8510 Cal Set, a limited instrument state is needed in order to return the instrument to the

same state that it was in when the calibration was done .

Three arrays of error correction data are de ned by using three D A T A statements . Some

instruments require these arrays to be in the proper order , from E1 to E3. In general, CITI le

implementations should strive to handle data arrays that are arranged in any order .

CITIFILE A.01.00

#NA VERSION HP8510B.05.00

NAME CAL_SET

#NA REGISTER 1

VAR FREQ MAG 4

DATA E[1] RI

DATA E[2] RI

DATA E[3] RI

#NA SWEEP_TIME 9.999987E-2

#NA POWER1 1.0E1

#NA POWER2 1.0E1

#NA PARAMS 2

#NA CAL_TYPE 3

#NA POWER_SLOPE 0.0E0

#NA SLOPE_MODE 0

#NA TRIM_SWEEP 0

#NA SWEEP_MODE 4

#NA LOWPASS_FLAG -1

#NA FREQ_INFO 1

#NA SPAN 1000000000 3000000000 4

#NA DUPLICATES 0

#NA ARB_SEG 1000000000 1000000000 1

#NA ARB_SEG 2000000000 3000000000 3

VAR_LIST_BEGIN

VAR_LIST_END

The CITI le Data Format and K eyword Reference A-5

Show 6 more labels / short text on this page
Example · 1000000000 · 2000000000 · 2500000000 · 3000000000 · Continued next page
Page 650 of the Agilent Technologies 8753D user manual — BEGIN
Page 650 · click to enlarge
Page 650

BEGIN

1.12134E-3,1.73103E-3

4.23145E-3,-5.36775E-3

-0.56815E-3,5.32650E-3

-1.85942E-3,-4.07981E-3

END

BEGIN

2.03895E-2,-0.82674E-2

-4.21371E-2,-0.24871E-2

0.21038E-2,-3.06778E-2

1.20315E-2,5.99861E-2

END

BEGIN

4.45404E-1,4.31518E-1

8.34777E-1,-1.33056E-1

-7.09137E-1,5.58410E-1

4.84252E-1,-8.07098E-1

END

When an instrument's frequency list mode is used, as it was in Example 4, a list of frequencies

is stored in the le after the V AR LIST BEGIN statement. The unsorted frequency list

segments used by this instrument to create the V AR LIST BEGIN data are de ned in the #NA

ARB SEG statements .

The descriptions and examples shown here demonstrate how CITI le may be used to store and

transfer both measurement information and data. The use of a single , common format will

allow data to be more easily moved between instruments and computers .

A-6 The CITI le Data Format and K eyword Reference

Example (continued) · Conclusion

Page 651 of the Agilent Technologies 8753D user manual — A - The CITIfile Data Format and Keyword Reference
Page 651 · click to enlarge
Page 651

The CITI le K eyword Reference

K eyword Explanation and Examples

CITIFILE CITIFILE A.01.01 identi es the le as a CITI le , and indicates the

revision level of the le . The CITIFILE keyword and revision code must

precede any other keywords . The CITIFILE keyword at the beginning

of the package assures the device reading the le that the data that

follows is in the CITI le format. The revision number allows for future

extensions of the CITI le standard. The revision code shown here

following the CITIFILE keyword indicates that the machine writing this

le is using the A.01.01 version of CITI le as de ned here . Any future

extensions of CITI le will increment the revision code .

NAME NAME CAL_SET allows the current CITI le \package" to be named. The

name of the package should be a single word with no embedded spaces .

A list of standard package names follows:

RA W D A T A Uncorrected data.

D A T A Data that has been error corrected. When only a

single data array exists , it should be named D A T A.

FORMA TTED Corrected and formatted data.

MEMORY Data trace stored for comparison purposes .

CAL SET Coecients used for error correction.

CAL KIT Description of the standards used

DELA Y T ABLE Delay coecients for calibration.

V AR VAR FREQ MAG 201 de nes the name of the independent variable

(FREQ), the format of values in a V AR LIST BEGIN table (MA G, if

used), and the number of data points (201). Typical names for the

independent variable are FREQ (in Hz), TIME (in seconds), and POWER

(in dBm). F or the V AR LIST BEGIN table , only the \MA G" format is

supported at this point.

# #NA POWER1 1.0E1 allows variables speci c to a particular type of

device to be de ned. The pound sign (#) tells the device reading the

le that the following variable is for a particular device . The \NA"

shown here indicates that the information is for a Network Analyzer .

This convention allows new devices to be de ned without fear of

con ict with keywords for previously de ned devices . The device

identi er (i.e . NA) may be any number of characters .

SEG LIST BEGIN SEG_LIST_BEGIN indicates that a list of segments for the independent

variable follow . F ormat for the segments is: [segment type] [start]

[stop] [number of points]. The current implementation only

supports a single segment. If there is more than one segment, the

V AR LIST BEGIN construct is used. CITI le revision A.01.00 supports

only the SEG (linear segment) segment type .

SEG LIST END SEG_LIST_END de nes the end of a list of independent variable

segments .

The CITI le Data Format and K eyword Reference A-7

Label De nition

Page 652 of the Agilent Technologies 8753D user manual — A - The CITIfile Data Format and Keyword Reference
Page 652 · click to enlarge
Page 652

V AR LIST BEGIN VAR_LIST_BEGIN indicates that a list of the values for the independent

variable (declared in the V AR statement) follow . Only the MA G format

is supported in revision A.01.00.

V AR LIST END VAR_LIST_END de nes the end of a list of values for the independent

variable .

D A T A DATA S[1,1] RI de nes the name of an array of data that will be read

later in the current CITI le package , and the format that the data will

be in. Multiple arrays of data are supported by using standard array

indexing. V ersions A.01.00 and A.01.01 of CITI le only support the

RI (real and imaginary) format, and a maximum of two array indexes .

Commonly used array names include the following:

\S" for \S parameter" ExampleS[2,1]
\E" for \Error term" ExampleE[1]
\USER" for \User parameter" ExampleUSER[1]

\V OL T A GE" Example: V OL T A GE[1]

\V OL T A GE RA TIO" for a ratio of two

ExampleV OL T A GE RA TIO[1,0]

voltages (A/R).

CONST ANT CONSTANT [name] [value] allows for the recording of values which

don't change when the independent variable changes .

CONST ANT s are part of the main CITI le de nition. Users must not

de ne their own CONST ANT s . Use the #KEYWORD device speci cation

to create your own KEYWORD instead. The #NA device speci cation is

an example of this .

No constants were de ned for revision A.01.00 of CITI le . CITI le

revision A.01.01 de ned the following constant:

CONST ANT TIME [year] [month] [day] [hour] [min] [secs]

Example:

COMMENT YEAR MONTH D A Y HOUR MINUTE SECONDS

CONST ANT TIME 1991 02 26 17 33 53.25

The COMMENT statement is not absolutely required, but is highly

recommended to aid readability .

The year should always be the full four digits (\1991" is correct, but

\91" is not). This is to avoid problems with the year 2000, when the

shortened version of the year will be \00."

The hour value should be in 24-hour \military" time .

When writing a CITI le and the fractional seconds value is zero , then

the \seconds" value may be printed either with or without a decimal

pointeither \47.0" or \47" would be acceptable . When reading

a CITI le , the seconds value should always be read as if it were a

oating point number .

A-8 The CITI le Data Format and K eyword Reference

Page 654 of the Agilent Technologies 8753D user manual — B
Page 654 · click to enlarge
Page 654

battery backup for memory, 12-2

measurement setup diagram, 3-31

bias inputs and fuses locations, 1-11

annotations of display, 1-7

blanking the display, 6-42

application and operation concepts, 6-1

broad band power meter calibration

ampli er testing, 6-145{148

C10 (pass control capabilities), 11-18

C1,C2,C3 (system controller capabilities),

raw data , 4-36

assign classes, 5-29, 5-31

C2 status notation, 1-7

A TN (attention) control line, 11-17

attenuation at mixer ports, 3-2

HP-IB, 11-5

allowing repetitive switching, 6-19

test port return, 11-2

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

AUTO-FEED , 4-12

calculate statistics of the measurement data,

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

automatic le naming convention for LIF

auto sweep time mode

how to set, 5-51

auxiliary input connector location, 1-11

for mixer measurements, 3-6

for non-insertable devices, 6-101

for noninsertable devices, 5-41

how to widen, 5-52

A VG

A VG

menu map, 8-2

A vg status notation, 1-7

B

power meter, 5-33, 6-95

fault location measurements, 6-123

procedure to modify a user kit, 6-76

re ection measurements, 6-119

response and isolation, 6-73

system:how to widen, 5-52

basic measurement sequence and example,

basic system maintenance service option,

basic talker (T6), 11-18

Show 65 more labels / short text on this page
analyzer description, 1-2 · analyzer display · battery life, 4-33 · bi-directional lines, 11-16 · anti-static mat, 7-36 · applications · connections, 3-13 · bus · mixer testing, 6-149{160 · device modes, 11-19 · on-wafer measurements, 2-76 · structure, 11-15, 11-16 · arrays · exibility, 4-36 · format, 6-7 · format , 4-36 · atmospheric conditions, 7-36 · cabinet, 11-4 · cabinet dimensions, 7-37 · cables · attenuator, 6-2 · interconnecting, 5-2 · interface, 11-5 · switch protection, 6-19 · AUTO-FEED use, 4-5 · calibrate menu, 6-73 · automated measurements, 6-138{144 · calibrating · noninsertable devices, 5-39 · format, 4-19 · calibration · concepts, 6-50{101 · considerations, 6-65 · data lost, 12-2 · available options, 1-12 · device measurements, 6-65 · averaging, 6-44 · fast 2-port, 5-54 · changes, 5-56 · sweep-to-sweep, 6-5 · averaging factor · full two-port, 6-73 · · key, 6-44 · in memory, 12-2 · · interpolated error-correction, 6-72 · measurement parameters, 6-65 · modifying kits, 6-76 · omitting isolation, 6-65 · · key, 6-10 · one-port, 6-73 · bandpass mode · receiver, 5-11 · transmission measurements, 6-121 · response, 6-73 · bandwidth · restarting, 6-75 · saving data, 6-65 · specifying kits, 6-75 · standard devices, 6-66 · standards, 5-6 · temperature, 12-5 · Index-2
Page 655 of the Agilent Technologies 8753D user manual — CAL
Page 655 · click to enlarge
Page 655

TRL*/LRM*, 6-85

channel display titling, 2-8

using on multiple analyzers, 12-5

verify performance of, 6-84

channel stimulus coupling, 6-20

characteristic power meter calibration speed

saving modi ed kit, 5-28

characterizing microwave systematic errors,

calibration factors entry, 5-34

characters that are valid, 11-22

label standard menu, 6-81

chop sweep mode, 6-23, 6-100

saving modi ed or user kit, 5-28

how to activate, 5-54

specify o set, 6-80

class label, 5-30, 5-32

standard de nitions, 6-78

cleaning of connectors, 2-2

clearing a sequence from memory, 2-57

clock setting, 4-30, 11-13

code naming conventions, 11-21

color printer use, 4-6

electrical delay of, 5-6

electrical o set, 6-67

frequency response of, 6-66

to key cross reference, 9-48

outline of modi cation, 5-26

CAL

cal kit menu, 6-75

menu map, 8-3

CAL

directional coupler response, 5-35

capabilities of the analyzer, 1-2

electrical delay of calibration standards,

care of connectors, 2-2

cascading multiple sequences, 2-61

complete service request capabilities (SR1),

completion time for measurements, 7-33

center frequency set with markers, 2-18

centronics (parallel) interface, 1-10

what you can save to , 4-34

changes between the HP 8753A/B/C/D, 1-13

change system bandwidth, 5-57

application and operation, 6-1

changing the sequence title, 2-58

channel 1 and 2 ratio measurement, 2-7

Show 67 more labels / short text on this page
TRL*/LRM* two-port, 6-74 · channel keys, 6-8 · uncoupling stimulus, 6-8 · validity of, 6-71 · channel power · coupling, 6-15 · calibration arrays · channel power coupling · memory requirements, 12-3 · harmonic operation, 6-115 · saving, 4-37 · calibration constants · channel viewing, 2-5 · entering, 5-26 · examining, 5-26 · and accuracy, 5-33 · modifying, 5-26 · characteristics, 7-1 · calibration kit · class assignments, 6-81 · choosing · de nitions, 6-76 · display elements, 4-12 · scale, 4-15 · modify menu, 6-77 · terminology, 6-76 · calibration kits, 11-1 · calibration sets · CMOS memory, 12-2 · using, 12-5 · calibration standards · de ning, 6-91 · colors · de nitions, 5-26 · modify menu, 6-42 · recalling, 6-42 · saving, 6-42 · fabricating, 6-91 · setting default, 6-42 · command · fringe capacitance, 6-67 · syntax, 11-21 · modifying, 5-26 · optimizing, 5-6 · compatible · plotters, 11-4 · · key, 6-71 · printers, 11-4 · compensating for · · on LCD, 10-1 · · key, 6-10 · · compression measurement, 2-45 · computer, 11-6 · centronics interface, 7-35 · controllers, 11-15 · concepts · accuracy enhancement, 6-50{101 · changing · active channel, 6-8 · parameters, 5-5 · calibration, 6-50{101 · connections, 6-161 · Index-3
Page 656 of the Agilent Technologies 8753D user manual — COPY
Page 656 · click to enlarge
Page 656

care quick reference, 2-2

for external monitor, 1-10 , 1-11

parallel (centronics) interface, 1-10

probe power source, 1-5

serial (RS-232) interface, 1-10

conditions for environment, 7-36

conditions for error-correction, 5-4

test set interconnect, 1-11

external disk drive, 11-12

plotter, 4-8 , 11-9

connectors on front panel, 7-33

connectors on rear panel, 7-34

a plot function, 4-8

considerations for mixer measurements, 3-2

a print function, 4-3

of disk le, 4-33

printer, 4-4 , 4-9, 11-8

of internal memory registers, 4-33

the analyzer for the printer, 4-4

continuous and discrete markers, 2-9

con guring the analyzer for the peripheral,

continuous correction mode for power meter

continuous mode , test set switch, 5-54

broad band power meter calibration, 3-13

conversion loss using the tuned receiver

controlling the test set switch, 5-54

down converter port, 3-3

conventions for code naming, 11-21

rst portion of conversion compression

conversion compression mixer measurement,

group delay measurement, 3-25

and output power as a function of input

high dynamic range swept IF conversion

power level example, 3-28

mixer isolation measurement, 3-34

mixer RF feedthrough measurement, 3-36

one-sweep power meter calibration for

using the frequency o set mode, 3-7

using the tuned receiver mode connections,

peripheral to analyzer, 4-8

printer to analyzer, 4-3

using tuned receiver mode, 3-17

Rchannel and source calibration, 3-8

menu map, 8-5

COPY

response calibration, 3-18, 3-34, 3-36

correcting for measurement errors, 6-71

second portion of conversion compression

up converter port, 3-3

Show 62 more labels / short text on this page
data processing, 6-4{7 · care of, 2-2 · entry keys, 6-9{10 · error-correction, 6-50{101 · external am, 1-11 · frequency domain, 6-117{137 · external trigger, 1-11 · instrument state, 6-102{108 · markers, 6-47{49 · for HP-IB, 1-10 · measurement calibration, 6-50{101 · for keyboard, 1-10 · response function, 6-25{46 · limit test, 1-11 · sequencing, 6-138{144 · stimulus function, 6-11{24 · system overview, 6-2{3 · R channel, 1-5 · time domain, 6-117{137 · test sequence, 1-11 · con guration · connector repeatability, 5-2 · connectors · rear panel, 1-10 · power meter, 11-12 · con guring · contents · interface port, 11-12 · 4-8, 11-8 · calibration, 5-37 · connecting peripherals, 11-7 · connections, 6-161 · controller · adapters, 6-161 · external, 11-6 · interface function, 11-15 · control lines, 11-16 · mode, 3-22 · conversion, 6-6 · measurement, 3-29 · xture design, 6-162 · xtures, 6-162 · conversion loss · loss measurement, 3-15 · equation, 3-11 · example measurement, 3-11 · procedure, 3-12 · mixer measurements, 3-9 · conversion menu, 6-27 · · receiver calibration, 3-14 · correction, 5-7 · measurement, 3-30 · changing parameters, 5-5 · stimulus state, 5-5 · connection techniques, 2-2 · types, 5-4 · connector · correction procedures · auxiliary input, 1-11 · use of, 5-5 · Index-4
Page 657 of the Agilent Technologies 8753D user manual — D
Page 657 · click to enlarge
Page 657

Cor status notation, 1-7

plotter pen numbers, 4-13

couple and uncouple display markers, 2-14

de ned plotting components, 4-12

de ne standard menus, 6-78

de nition of front panel keys, 9-2

creating a sequence, 2-54

cross reference of key function to HP-IB

all instrument state les, 4-41

uncorrected speci cations, 7-2

an instrument state, 4-41

C? status notation, 1-7

a single instrument state le, 4-41

CW time sweep, 6-23

CW time-to-frequency domain, 6-134

D

Del status notation, 1-7

delta () markers, 2-11

description and options, 1-1

description of hardkeys and softkeys, 9-2

divided by memory, 2-7

loss of power calibration, 5-33

deviation from linear phase measurement,

device clear (DC1), 11-18

device types for HP-IB, 11-15

trace saved to the display memory, 2-6

diagram of measurement frequencies, 3-10

port extensions and electrical delay, 5-3

data and memory viewing, 2-6

the HP 8753A/B/C/D, 1-13

dimensions of 7 mm test port, 7-33

dimensions of analyzer, 7-37

de nition of, 6-5

directional coupler compensation, 5-35

data processing ow diagram, 4-37

uncorrected speci cations, 7-2

for internal memory, 4-33

discrete and continuous markers, 2-9

DC1 (complete device clear), 11-18

the sweep rate, 5-48

storage of a sequence, 2-59

the time delay, 5-49

line types for plotter, 4-14

con guring to plot, 4-11

pen numbers for plot elements, 4-13

using with other models, 12-5

dispersion e ects, 6-94

Show 62 more labels / short text on this page
coupled · channels, 2-6, 2-7 · de ning · markers, 6-8 · a plot, 4-12 · coupling · the print, 4-5 · channel power, 6-15 · de nition of · channel stimulus, 6-20 · calibration standards, 5-26 · port power, 6-15 · power, 6-15 · delay · electrical, 6-38 · delay block · command, 9-48 · electrical, 6-6 · crosstalk, 6-53 · deleting · reducing, 5-57 · frequency segments, 5-35 · limit segments, 2-42 · sequence commands, 2-56 · damage level, 7-28 · data · bus, 11-16 · designing xtures, 6-162 · processing, 6-4{7 · rate, 11-17 · statistics, 2-26 · transfer, 11-16 · di erences between · units, 11-23 · data arrays · exibility, 4-36 · data point · DIN keyboard, 7-36 · directivity, 6-51 · data retention · directory size, 12-4 · data storage · types of, 12-1 · disk · external, 12-4 · debug mode, 11-23 · le contents, 4-33 · decrease · formatting, 4-43 · frequency span, 5-50 · internal, 12-4 · storage problems, 4-43 · default · storing data, 12-4 · le names, 4-33 · disk drive · location, 1-4 · plotting parameters, 4-16 · disk les · printing parameters, 4-6 · default settings, 12-6 · Index-5
Page 658 of the Agilent Technologies 8753D user manual — TRL*, 5-23
Page 658 · click to enlarge
Page 658

E

E2 (tri-state drivers), 11-18

displayed measurement title, 4-29

of calibration standards, 5-6

electrical length and phase distortion

causes of measurement problems, 5-48

how to improve measurement results, 5-48

making accurate measurements of, 5-48

electrical o set, 6-67

SWR, 6-33

end or identify (EOI) control line, 11-17

display functions, 2-5, 6-39

ensuring valid calibration, 6-71

entering calibration constants, 5-26

entry block location, 1-5

DISPLA Y

DISPLA Y

menu map, 8-6

ENTRY OFF

divide measurement data by the memory

EOI (end or identify) control line, 11-17

error-correction, 2-4, 5-2, 5-7

does not respond to parallel poll (PPO), 11-18

characteristics a ected, 5-4

down converter port connections, 3-3

full two-port calibration, 6-73

full two-port measurements, 5-20

DT1 (responds to a group execute trigger),

dual channel mode, 2-5, 6-40

one-port re ection measurements, 5-17

response and isolation, 6-73

response and isolation for re ection

duration of measurements, 7-33

response and isolation for transmission

response for re ection measurements, 5-8

response for transmission measurements,

TRL*, 5-23

TRL*/LRM* two-port calibration, 6-74

Show 74 more labels / short text on this page
display · blanking, 6-42 · format, 6-29 · edit · information, 1-6 · list menu, 6-23 · location, 1-4 · subsweep menu, 6-24 · markers, 2-9 · editing · markers activation, 2-10 · a sequence, 2-56 · memory, 6-7 · frequency segments, 5-34 · memory trace, 2-6 · limits, 6-107 · of analyzer, 1-6 · limit segments, 2-42 · setting intensity, 6-41 · segments, 6-107 · status notations, 1-7 · test sequences, 6-139 · titling, 2-8 · electrical · trace math, 2-7 · delay block, 6-6 · electrical delay, 6-38 · display format · group delay, 6-30 · imaginary, 6-34 · measurement, 2-30 · linear magnitude, 6-33 · electrically long devices · log magnitude, 6-29 · phase, 6-30 · polar, 6-32 · real, 6-34 · smith chart, 6-31 · electrostatic precautions, 7-36 · color, 6-41 · dual channel, 6-40 · · key, 6-39 · entry keys, 6-9{10 · · · key, 6-10 · environmental characteristics, 7-36 · trace, 2-7 · documents · reference, 6-163 · calibration standards, 5-6 · changing parameters, 5-5 · DOS format, 11-5 · concepts, 6-50{101 · drift · conditions, 5-4 · frequency, 5-3 · temperature, 5-2 · interpolated, 5-6 · invalidating, 5-5 · dual channel, 6-8 · one-port calibration, 6-73 · problems with, 6-40 · dual trace, 6-8 · measurements, 5-13 · dynamic accuracy, 7-31 · speci cations, 7-29 · measurements, 5-15 · dynamic range, 7-1 · response calibration, 6-73 · increasing, 5-56 · limitations, 7-1 · Index-6
Page 659 of the Agilent Technologies 8753D user manual — F ORMA T
Page 659 · click to enlarge
Page 659

TRM*, 5-23

CW frequency range, 6-111

external trigger circuit, 7-35

external trigger (EXT TRIGGER), 7-34

external video monitor, 7-35

ext mon. connector, 1-10, 1-11

examining calibration constants, 5-26

ext status notation, 1-7

F

conversion loss and output power as a

function of input power level, 3-28

conversion loss measurement, 3-11

fast 2-port calibration, 5-54

xed IF mixer measurement, 3-23

faster sweep speed, 5-50

group delay measurement, 3-26

mixer LO to RF isolation measurement,

using low pass mode, 6-123

feature di erences between

mixer RF feedthrough measurement, 3-37

HP 8753A/B/C/D, 1-13

plot quadrants, 4-17, 4-28

plotting two les on the same page, 4-27

printing two measurements, 4-7

features added to the analyzer, 1-2

signal ow in a mixer, 3-33

features of front panel, 1-4

spectrum of RF ,LO ,and IF signals present

in a conversion loss measurement, 3-7

RF, 3-35

swept IF conversion loss measurement,

swept power conversion compression

how to re-save, 4-41

viewing both channels with a split display,

extended listener capabilities (LEO), 11-18

internal memory registers, 4-33

am connector location, 1-11

trigger connector location, 1-11

ltering for mixers, 3-2

external AM input (EXT AM), 7-34

rst portion of conversion compression

external auxiliary input (A UX INPUT), 7-34

xed IF mixer measurement

xed IF mixer measurements, 3-17

how to solve problems with, 4-43

ag markers for gating, 2-70

external monitor connector, 1-10

at limit lines, 2-37

external reference frequency input (EXT

oppy disk le contents, 4-33

REF INPUT), 7-34

external source mode, 6-109

area of display, 1-8

F ORMA T

Show 55 more labels / short text on this page
types, 5-4 · description, 6-110 · error messages, 10-2 · frequency modulation, 6-111 · alphabetically listed, 10-2{25 · manual, 6-110 · numerically listed, 10-26{31 · requirements, 6-111 · error model · speci cations, 7-29 · device measurement, 6-59 · sweep types, 6-111 · one-port, 6-53 · test setup, 6-110 · two-port, 6-59 · errors · removing, 5-4 · ESD precautions, 7-36 · example · fan location, 1-11 · fault location measurements · features · rear panel, 1-10 · feedthrough · le · deleting, 4-41 · measurement, 3-32 · recalling, 4-42 · renaming, 4-42 · saving, 4-33 · example procedures, 2-1 · le contents · oppy disk, 4-33 · external · le names, 12-4 · monitor connector, 1-11 · default, 4-33 · measurement connections, 3-29 · external controller, 11-6 · external disk, 12-4 · example, 3-23 · external disk drive · con guration, 11-12 · xed markers, 2-11 · xture design, 6-162 · external keyboard, 7-36 · format, 6-7 · applications, 6-109 · arrays, 6-7 · auto, 6-110 · menu, 6-29 · capture range, 6-111 · · key, 6-29 · Index-7
Page 660 of the Agilent Technologies 8753D user manual — H
Page 660 · click to enlarge
Page 660

G

F ORMA T

menu map, 8-6

gain and reverse isolation measurement,

NNNNNNNNNNNNNNNNNNNNNNNNNN

FORMFEED , 4-29

gain compression measurement, 2-45

demodulating the results, 6-135

gate span setting, 2-70

response and isolation error-corrections,

selection for mixers, 3-2

gating e ects in frequency domain, 2-72

Gat status notation, 1-7

transform from CW time, 6-134

GPIO mode, 6-105, 6-139

frequency o set mode, 3-7

frequency o set operation, 6-112

compatible instrument modes, 6-113

group delay accuracy vs . aperture, 7-31

group delay or phase mixer measurements,

group execute trigger response (DT1), 11-18

guidelines for code naming, 11-21

H

o set frequency, 6-113

receiver and source requirements, 6-114

H=2 status notation, 1-8

H=3 status notation, 1-8

hardkey de nitions, 9-2

harmonic mode option, 1-12

coupling channel power, 6-115

frequency response of calibration standards,

front panel access key for softkeys, 9-69

front panel connectors, 7-33

front panel features, 1-4

front-panel key de nitions, 9-2

high dynamic range swept IF conversion loss

full-acceptor handshake (AH1), 11-18

full-source handshake (SH1), 11-18

Show 69 more labels / short text on this page
format markers · calibration, 6-73 · polar, 2-14 · error-correction, 5-20 · Smith, 2-15 · · formatting · a disk, 4-43 · forward transform, 6-135 · gate · ags, 2-70 · range, 6-136 · parameter setting, 2-70 · frequency · shaping, 2-70 · drift, 5-3 · gating, 6-6 · procedure, 2-70 · time domain, 6-133 · speci cations, 7-27 · frequency domain · concepts, 6-117{137 · G+jB MKR, 2-16 · GPIO interface, 1-10 · frequency o set · speci cations, 7-29 · group delay, 6-35 · format, 6-30 · mixer testing, 6-149 · measurement, 2-33 · operation, 3-4 · measurement connections, 3-25 · measurement example, 3-26 · applications, 6-112 · mixer measurements, 3-24 · speci cations, 7-31 · description, 6-113 · display annotations, 6-114 · error message, 6-114 · frequency hierarchy, 6-113 · frequency ranges, 6-113 · LO frequency, 6-113 · receiver frequency, 6-113 · spurious signals, 6-114 · handshake lines, 11-16 · test setup, 6-112 · frequency range · de ning, 6-11 · harmonic operation, 6-115 · frequency response, 6-53 · accuracy, 6-116 · error-corrections, 5-8 · dual-channel, 6-115 · frequency range, 6-116 · frequency span · input power, 6-116 · decrease, 5-50 · single-channel, 6-115 · frequency-to-time domain, 6-117 · test setup, 6-115 · fringe capacitance, 6-67 · harmonics · speci cations, 7-27 · harmonics measurement, 2-66 · harmonics (option 002) · speci cations, 7-28 · measurement connections, 3-15 · full two-port · Index-8
Page 661 of the Agilent Technologies 8753D user manual — Index
Page 661 · click to enlarge
Page 661

high dynamic range swept RF/IF conversion

error-correct for frequency response, 5-8

error-correct for frequency response and

high-stability frequency reference(10 MHz)

error-correct for full two-port

high stability frequency reference option,

error-correct for one-port re ection

Hld status notation, 1-8

hold mode , test set switch, 5-54

error-correct for response and isolation

for re ection measurements, 5-13

abort a print or plot process, 4-30

error-correct for response and isolation

activate a xed marker, 2-11

for transmission measurements, 5-15

activate chop sweep mode, 5-54

error-correct for TRL*, 5-23

activate display markers, 2-10

error-correct for TRM*, 5-23

calculate data statistics, 2-26

calibrate the analyzer receiver to measure

error-correct response for re ection

change measurement averaging, 5-56

error-correct response for transmission

change system bandwidth, 5-56

change the sequence title, 2-58

format a disk, 4-43

gate out responses, 2-70

clear a sequence from memory, 2-57

increase sweep speed, 5-50

compensate for directional coupler

insert a sequence command, 2-56

load a sequence from disk, 2-60

con gure a plot function, 4-8

make a basic measurement, 2-3

con gure a plot function to disk drive,

make accurate measurements of electrically

con gure a plot function using and HPGL/2

measure conversion loss using tuned

con gure a plot function using a pen

measure deviation from linear phase, 2-32

measure devices using the time domain,

con gure a print function, 4-3

construct a loop structure in a sequence,

measure electrical length and phase

control the test set switch, 5-54

measure xed IF mixers, 3-17

couple and uncouple display markers, 2-14

measure gain compression, 2-45

create a limit test sequence, 2-64

measure group delay, 2-33

create a test sequence, 2-54

measure group delay or phase for mixers,

create at limit lines, 2-37

create single point limits, 2-41

measure high dynamic range conversion

create sloping limit lines, 2-39

decrease frequency span, 5-50

measure magnitude and insertion phase

de ne line types, 4-14

de ne the plot, 4-12

measure mixer isolation, 3-33

de ne the print, 4-5

measure mixer LO to RF isolation, 3-33

delete all instrument state les, 4-41

delete an instrument state, 4-41

measure re ection response in time domain,

delete a single instrument state le, 4-41

delete limit segments, 2-42

measure reverse isolation and gain, 2-49

divide measurement data by the memory

measure swept harmonics, 2-66

measure swept RF/IF mixers, 3-7

edit a sequence, 2-56

measure transmission response in time

edit limit segments, 2-42

edit power calibration factors, 5-34

modify a sequence command, 2-57

enter the power sensor calibration data,

modify calibration standards, 5-26

modify TRL calibration standards, 5-28

Show 23 more labels / short text on this page
loss, 3-12 · isolation, 5-13 · (option 001), 7-34 · measurements, 5-20 · measurements, 5-17 · how to · error-correct measurements, 5-7 · absolute power, 5-38 · measurements, 5-8 · measurements, 5-10 · clean connectors, 2-2 · response, 5-35 · long devices, 5-48 · compatible printer, 4-8 · receiver mode, 3-17 · plotter, 4-10, 11-11 · distortion, 2-30 · loss, 3-12 · response, 2-27 · measure mixers, 3-1 · trace, 2-7 · domain, 2-68 · Index-9
Page 662 of the Agilent Technologies 8753D user manual — HP-IB, 11-14
Page 662 · click to enlarge
Page 662

modify TRM calibration standards, 5-30

o set limit lines, 2-44

set auto sweep time mode, 5-51

output a single page of values, 4-30

set center frequency with markers, 2-18

output multiple plots to a single page using

set CW frequency using markers, 2-21

set display reference using markers, 2-20

output plot les, 4-20

set electrical delay using markers, 2-21

output plot les from a PC to a plotter,

set frequency range, 2-3

set frequency span with markers, 2-19

output plot les from a PC to a printer,

set source power, 2-4

set start frequency with markers, 2-17

output single page plots using a printer,

set stop frequency with markers, 2-17

set the measurement parameters, 2-3

output the entire list of values, 4-31

set the measurement type, 2-4

plot a measurement to disk, 4-19

set the sweep type, 5-53

plot measurements in page quadrants,

solve printing or plotting problems, 4-32

solve problems with disk storage, 4-43

plot multiple measurements on a full page,

stop a sequence, 2-55

store a sequence on a disk, 2-59

plot multiple measurements per page,

subtract the memory trace from the

measurement data trace, 2-7

plot one measurement per page, 4-16

title the active channel display, 2-8

plot to an HPGL compatible printer, 4-18

title the displayed measurement, 4-29

power meter calibrate, 3-6

print a sequence, 2-60

use analyzer display markers, 2-9

print multiple measurements per page,

use continuous and discrete markers, 2-9

use delta () markers, 2-11

print one measurement per page, 4-6

use frequency o set mode, 3-7

print or plot list values or operating

use limit lines, 2-36

NNNNNNNNNNNNNNNNNNNNNNNNNN

MKR ZERO to activate a xed reference

purge a sequence from disk, 2-60

ratio measurements in channel 1 and 2,

use polar format markers, 2-14

use receiver calibration, 5-11

recall a le, 4-42

use Smith chart markers, 2-15

reduce receiver crosstalk, 5-57

use the tuned receiver mode, 2-51

reduce receiver noise oor, 5-56

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REF=FIXED MKR key to activate

reduce the averaging factor, 5-52

a xed reference marker, 2-12

reduce the number of points, 5-52

view a single measurement channel, 5-53

reduce trace noise, 5-57

view both measurement channels, 2-5

rename a le, 4-42

view plot les on a PC, 4-20

re-save a le, 4-41

view the measurement data and memory

reset plotting parameters to default values,

widen system bandwidth, 5-52

reset the printing parameters to default

HP 8753A/B/C/D di erences, 1-13

HPGL

review limit line segments, 2-43

run a limit test, 2-43

outputting plot les to compatible printer

run a test sequence, 2-55

save a data trace to the display memory,

test le commands, 4-24

HPGL/2 compatible printer , con guring to

save an instrument state, 4-35

save measurement results, 4-36

HP-IB, 11-14

search for a target amplitude, 2-24

search for bandwidth, 2-25

search for maximum amplitude, 2-22

bus structure, 11-15, 11-16

search for minimum amplitude, 2-23

Show 16 more labels / short text on this page
sequence tests, 2-53 · a printer, 4-25 · 4-17, 4-26 · track amplitude, 2-25 · parameters, 4-30 · use · marker, 2-13 · use the · trace, 2-6 · values, 4-6 · initialization commands, 4-23 · , 4-23 · plot, 4-8 · address capability, 11-17 · addresses, 6-104, 11-20 · Index-10
Page 663 of the Agilent Technologies 8753D user manual — MEAS
Page 663 · click to enlarge
Page 663

instrument state function block location, 1-5

instrument state in memory, 12-1

instrument states in memory, 12-2

interconnect devices ,max.number, 11-17

message transfer scheme, 11-17

interconnection cable ,max. length, 11-17

pass control mode, 6-104

GPIO, 1-10

status indicators, 6-104, 11-19

system controller mode, 6-104

RS-232, 1-10

interface clear (IFC) control line, 11-16

HP-IB-command to key-function cross

interface function codes, 7-33

IF

how de ned for mixers, 3-2

how to widen, 5-52

IFC (interface clear) control line, 11-16

interpolated error-correction, 5-6, 6-72

measurements for mixers, 3-33

(reverse) and gain measurement, 2-49

test port power, 5-56

K

increasing measurement accuracy, 5-2

key de nitions, 9-1

key menu maps, 8-1

DOS, 11-5

LIF, 11-5

input ports menu, 6-28

A V G

inserting a sequence command, 2-56

insertion phase and magnitude response

CAL

DISPLA Y

frequency o set, 6-112

ENTRY OFF

F ORMA T

LOCAL

, 6-26, 6-27

MEAS

MENU

deleting all les, 4-41

deleting a single le, 4-41

Show 86 more labels / short text on this page
cables, 11-5 · memory requirements, 12-3 · connector, 1-10 · saving, 4-33, 4-35 · data rate, 11-17 · debug mode, 11-23 · device types, 11-15 · menu, 6-103 · interconnecting cables, 5-2 · multiple-controller capability, 11-17 · operation, 11-15 · interface · operational capabilities, 11-18 · addresses, 11-20 · cables, 11-5 · requirements, 11-17 · parallel, 1-10, 7-35 · talker/listener mode, 6-104 · reference, 9-48 · interface functions · HP-IB connector, 7-35 · controller, 11-15 · humidity conditions, 7-36 · listener, 11-15 · talker, 11-15 · interface port · con guration, 11-12 · internal memory · IF bandwidth, 6-45 · data retention, 4-33 · register contents, 4-33 · interpolated calibration · IF detection, 6-5 · restoring, 12-5 · imaginary format, 6-34 · impedance · isolation, 6-53 · speci cations, 7-28 · calibration, 6-73 · impedance conversion, 11-2 · increase · increasing · dynamic range, 5-56 · keyboard connector, 1-10 · sweep speed, 5-50 · keyboard (DIN), 7-36 · keyboards, 11-6 · information messages, 10-2 · initializing · keys · active channel, 6-8 · · , 6-44 · · , 6-10 · · , 6-71 · · , 6-10 · measurement, 2-27 · · · , 6-39 · instrument modes, 6-109{116 · external source, 6-109 · entry, 6-9 {10 · · , 6-10 · · , 6-29 · harmonic, 6-115 · network analyzer, 6-109 · instrument state, 6-102{108 · tuned receiver, 6-111 · · , 6-103 · instrument state, 6-102{108 · · contents, 4-33 · · , 6-12 · · · , 6-10 · response function, 6-25{46 · Index-11
Page 664 of the Agilent Technologies 8753D user manual — LOG MKR, 2-15, 2-16
Page 664 · click to enlarge
Page 664

line types available, 4-14

line voltage selector switch, 1-10

SYSTEM

LIN MKR, 2-15, 2-16

listener interface function, 11-15

listen mode (L), 11-19

keys referenced to front panel access key,

key to HP-IB command, 9-48

list frequency sweep, 6-22

kit label, 5-30, 5-32

printing or plotting, 4-30

RPG, 6-10

L (listen mode), 11-19

loading a sequence from disk, 2-60

label class menu, 6-83

uncorrected speci cations, 7-2

label kit, 5-30, 5-32

LOCAL

LOCAL

menu map, 8-7

label the classes, 5-30, 5-32

LCD, 1-6

10 MHz precision reference output, 1-11

LE0 (no extended listener capabilities), 11-18

10 MHz reference adjust, 1-11

active channel keys, 1-4

auxiliary input connector, 1-11

bias inputs and fuses, 1-11

centronics (parallel) interface, 1-10

external trigger connector, 1-11

instrument state function block, 1-5

limit test connector, 1-11

creating at limit lines, 2-37

creating single point limits, 2-41

line voltage selector switch, 1-10

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

creating sloping limit lines, 2-39

P1 and P2 in

SCALE PLOT [GRAT] mode,

deleting limit segments, 2-42

editing limit segments, 2-42

parallel (centronics) interface, 1-10

o setting limit lines, 2-44

port 1 and port 2, 1-5

reviewing limit line segments, 2-43

running a limit test, 2-43

probe power source connector, 1-5

R channel connectors, 1-5

limit test output (LIMIT TEST), 7-35

response function block, 1-4

RS-232 (serial) interface, 1-10

phase deviation measurement, 2-32

serial number plate, 1-10

linear frequency sweep, 6-21

serial (RS-232) interface, 1-10

linear magnitude format, 6-33

lines for control, 11-16

stimulus function block, 1-4

lines for handshaking, 11-16

test sequence connector, 1-11

line switch location, 1-4

test set interconnect, 1-11

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

LINE TYPE DATA , 4-14

locations of softkeys, 9-69

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

logarithmic frequency sweep, 6-22

LINE TYPE MEMORY , 4-14

log magnitude format, 6-29

LOG MKR, 2-15, 2-16

loop counter sequence, 2-62

Show 48 more labels / short text on this page
step, 6-10 · stimulus function, 6-11{24 · · , 6-109 · units terminator, 6-9 · · · · , 6-10 · list frequencies · modifying, 6-23 · list values · knob · load match, 6-52 · · key, 6-103 · menu, 6-83 · · location · LIF format, 11-5 · limit lines, 6-106 · analyzer display, 1-4 · limits · editing, 6-107 · o set, 6-108 · limit segments · disk drive, 1-4 · editing, 6-107 · entry block, 1-5 · limit test · connector location, 1-11 · fan, 1-11 · example sequence, 2-64 · fuse, 1-10 · limit testing, 2-36 · activating, 2-43 · line switch, 1-4 · pass/fail message, 1-9 · pass/fail indicators, 2-44 · preset key, 1-5 · sequencing, 2-64 · linear · softkey labels, 1-9 · line power, 7-36 · softkeys, 1-4 · line types · plotter, 4-14 · Index-12
Page 665 of the Agilent Technologies 8753D user manual — LOG MKR, 2-15, 2-16
Page 665 · click to enlarge
Page 665

loss of power meter calibration data, 5-33

setting center frequency, 2-18

LO to RF isolation for mixers, 3-33

setting start frequency, 2-17

setting stop frequency, 2-17

re ection measurements, 6-123

setting frequency range, 6-122

Smith chart markers, 2-15

Smith or polar format, 2-12

M

magnitude and insertion phase response

bandwidth on display, 1-9

electrically long devices, 5-48

man status notation, 1-8

using to calibrate non-insertable devices,

values area on display, 1-8

using to calibrate noninsertable devices,

MARKER F CTN

menu map, 8-8

MARKER

menu map, 8-8

math with data traces, 2-7

maximum amplitude search, 2-22

MEAS

key, 6-26, 6-27

active and inactive, 2-10

MEAS

menu map, 8-9

calculate data statistics, 2-26

continuous and discrete, 2-9

deviation from linear phase, 2-32

couple and uncouple display markers, 2-14

device under test, 2-4

electrical length and phase distortion, 2-30

gain and reverse isolation, 2-49

magnitude and insertion phase response,

functions on display, 2-9

LIN MKR, 2-15, 2-16

transmission response in time domain,

LOG MKR, 2-15, 2-16

using limit lines, 2-36

using the time domain, 2-68

polar format, 2-14, 6-49

measured inputs of display, 1-8

Re/Im MKR, 2-15, 2-16

conversion loss example, 3-11

data points:setting, 2-4

search for bandwidth, 2-25

search for maximum amplitude, 2-22

search for minimum amplitude, 2-23

search for target amplitude, 2-24

parameters set with markers, 2-16

set CW frequency, 2-21

set display reference, 2-20

type:setting, 2-4

set electrical delay, 2-21

set frequency span, 2-19

set measurement parameters, 2-16

Show 56 more labels / short text on this page
low pass mode · smith chart, 6-49 · transmission measurements, 6-125 · target mode, 6-49 · tracking amplitude, 2-25 · magnitude · using, 6-47{49 · speci cations, 7-29 · zero, 2-13 · marker stats · measurement, 2-27 · making accurate measurements · masking · time domain, 6-127 · making measurements, 2-1 · mass storage, 11-5 · matched adapters · marker · control, 2-10 · · · markers · activating, 2-10 · · · measure · coupled, 6-8 · delta mode, 6-48 · delta (), 2-11 · gain compression, 2-45 · xed, 2-11 · group delay, 2-33 · xed reference, 6-48 · increased accuracy, 5-2 · functions, 6-49 · G+jB MKR, 2-16 · swept harmonics, 2-66 · menu, 6-48 · modes, 6-49 · reference, 2-11 · measurement · channel viewing, 2-5 · relative mode, 2-11 · R+jX MKR, 2-16 · data statistics, 2-26 · error-correction, 5-4 · example, 2-3 · examples, 2-1 · searching, 2-22 · output results, 2-4 · search mode, 6-49 · sequence, 2-3 · measurement accuracy · calibration standards, 5-6 · changing parameters, 5-5 · Index-13
Page 666 of the Agilent Technologies 8753D user manual — F ORMA T
Page 666 · click to enlarge
Page 666

compensating for directional coupler

using the tuned receiver mode, 2-51

measurement setup diagram shown on

measurement setup from display, 3-10

performance veri cation, 5-3

3.5 mm test ports, 7-8

reference plane and port extensions, 5-3

75 type-F test ports, 7-14, 7-17, 7-20,

75 type-N test ports, 7-11

7 mm test ports, 7-2

measurement channel, how to view a single,

CMOS, 12-2

saving calibration data, 12-2

saving current instrument state, 12-1

measurement frequencies , diagram, 3-10

saving instrument states, 12-2

how to reduce, 5-52

measurement port speci cations, 7-2

uncorrected for errors, 7-2

memory math functions, 6-41

electrically long devices, 5-48

conversion loss for mixers, 3-7

conversion loss using tuned receiver mode,

HP-IB, 6-103

high dynamic range conversion loss for

LO to RF isolation for mixers, 3-33

mixer conversion compression, 3-28

mixer group delay or phase, 3-24

mixers xed IF, 3-17

MENU

A VG

CAL

power meter calibration for, 3-6

COPY

RF feedthrough for mixers, 3-35

DISPLA Y

swept RF/IF mixers, 3-7

F ORMA T

Show 82 more labels / short text on this page
response, 5-35 · connector repeatability, 5-2 · analyzer display, 3-31 · error-correction, 5-4 · frequency drift, 5-3 · measurement times, 7-33 · increasing, 5-2 · measurement title, 4-29 · interconnecting cables, 5-2 · measurement uncertainties · 50 type-N, 7-5 · temperature drift, 5-2 · measurement averaging · changing, 5-56 · measurement calibration · concepts, 6-50{101 · memory · available, 12-2 · battery backup, 12-2 · measurement errors · causes of, 6-51 · conserving, 12-5 · correcting for, 6-71 · display, 6-7 · crosstalk, 6-53 · external disk, 12-4 · directivity, 6-51 · insucient, 12-3 · frequency response, 6-53 · internal disk, 12-4 · isolation, 6-53 · non-volatile, 12-2 · load match, 6-52 · requirements, 12-3 · source match, 6-52 · tracking, 6-53 · measurement points · types of, 12-1 · volatile, 12-1 · memory-intensive operations, 12-5 · measurement problems · memory register · contents, 4-33 · measurement programs, 11-6 · memory trace, 2-6 · measurement results · memory trace arrays · saving, 4-36 · memory requirements, 12-3 · measurements · menu · calibrate, 6-73 · cal kit, 6-75 · input ports, 6-28 · mixers, 3-12 · limits, 6-106 · power, 6-13 · mixer, 3-1 · scale reference, 6-38 · mixer considerations, 3-2 · S-parameter, 6-27 · sweep types, 6-21 · system, 6-106 · mixer isolation, 3-33 · trigger, 6-18 · · key, 6-12 · non-coaxial, 2-76 · menu map, 8-1 · optimized, 5-1 · · , 8-2 · optimizing, 4-43 · · , 8-3 · · , 8-5 · · , 8-6 · · , 8-6 · Index-14
Page 667 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNN
Page 667 · click to enlarge
Page 667

LOCAL

MARKER

MARKER F CTN

amplitude and phase tracking, 6-159

MEAS

MENU

PRESET

SA VE/RECALL

SCALE REF

SEQ

SYSTEM

frequency o set mode, 6-149

message area of display, 1-8

message transfer scheme, 11-17

methods of HP-IB operation, 11-15

microwave connector care, 2-2

minimum loss pad, 11-2

SWR, 6-157

tuned receiver mode, 6-149

NNNNNNNNNNNNNNNNNNNNNNNNNN

isolation measurement connections, 3-34

MKR ZERO example procedure, 2-13

LO to RE isolation measurement example,

auto sweep time, 5-51, 6-16

RF feedthrough measurement connections,

frequency o set, 3-7

manual sweep time, 6-16

RF feedthrough measurement example,

tuned receiver , 3-17

amplitude and phase tracking, 3-27

frequency o set, 6-112

attenuation at ports, 3-2

conversion loss using frequency o set, 3-7

conversion loss using tuned receiver mode,

eliminating unwanted signals, 3-2

modes for bus device, 11-19

frequency o set mode operation, 3-4

modi ed user kit, 5-28

group delay or phase measurement, 3-24

calibration kit menu, 6-77

high dynamic range conversion loss, 3-12

the cal kit thru de nition, 5-40

how RF and IF are de ned, 3-2

LO to RF isolation, 3-33

a sequence command, 2-57

minimizing source and load mismatches,

power meter calibration, 3-6

reducing the e ect of spurious responses,

TRL calibration standards, 5-28

TRM calibration standards, 5-30

swept RF/IF measurement, 3-7

Show 77 more labels / short text on this page
· , 8-7 · mixer testing, 6-149{160 · · , 8-8 · accuracy, 6-150 · · , 8-8 · · , 8-9 · attenuation, 6-151 · · , 8-10 · conversion compression, 6-158 · · , 8-12 · conversion loss, 6-156 · · , 8-11 · down conversion, 6-153 · · , 8-12 · ltering, 6-152 · · , 8-13 · frequency accuracy, 6-153 · · , 8-15 · frequency selection, 6-153 · messages · frequency stability, 6-153 · error, 10-1 · group delay, 6-159 · information, 10-2 · isolation, 6-156 · LO feedthru, 6-156 · parameters, 6-150 · microprocessor, 6-3 · phase linearity, 6-159 · phase measurements, 6-158 · minimum · return loss, 6-157 · amplitude search, 2-23 · RF feedthru, 6-157 · · · key, 6-10 · mixer · up conversion, 6-153 · mode · modes · mixer measurements, 3-1 · external source, 6-109 · mixers · forward transform, 6-118 · harmonic, 6-115 · conversion compression, 3-28 · instrument, 6-109{116 · network analyzer, 6-109 · pass-control, 11-20 · system-controller, 11-19 · talker/listener, 11-20 · ltering, 3-2 · tuned receiver, 6-111 · xed IF, 3-17 · frequency selection, 3-2 · modify · modifying · isolation measurements, 3-33 · a sequence, 2-56 · calibration constants, 5-26 · calibration kits, 6-76 · calibration standards, 5-26 · list frequencies, 6-23 · RF feedthrough, 3-35 · monitor, 7-35 · Index-15
Page 668 of the Agilent Technologies 8753D user manual — O
Page 668 · click to enlarge
Page 668

of the network analyzer, 6-2

operational capabilities for HP-IB, 11-18

operation of HP-IB, 11-15

multiple sequence cascading, 2-61

N

network analyzer mode, 6-109

new features to the analyzer, 1-2

002, harmonic mode, 1-12

NNNNNNNNNNNNNNNNNNNNNNN

NEWLINE , 4-29

006, 6 GHz operation, 1-12

no extended talker capabilities (TEO), 11-18

010, time domain, 1-12

011, receiver con guration, 1-12

+02A, standard system maintenance

noise level speci cations, 7-28

+02B , basic system maintenance service,

075, 75 impedance, 1-12

1CM, rack mount ange kit without handles,

1CP , rack mount ange kit with handles,

1D5, high stability frequency reference,

non-operating storage conditions, 7-36

+22A, return to HP full service agreement,

22B , return to HP repair agreement, 1-13

+22C, return to HP calibration agreement,

number of HP-IB devices allowed, 11-15

number of listeners allowed, 11-15

+22G, return to HP calibration, 1-13

+23G, on-site system veri cation, 1-13

how to reduce, 5-52

service and support, 1-13

number of sweeps , test set switch, 5-54

O

options and description, 1-1

o setting limit lines, 2-44

Ofs status notation, 1-7

Of? status notation, 1-7

one-sweep power meter calibration for mixer

multiple plots to a single page using a

on-site system veri cation option, 1-13

single page plots using a printer, 4-24

from a PC to a plotter, 4-22

printing or plotting, 4-30

from a PC to a printer, 4-23

frequency o set, 6-112

Show 58 more labels / short text on this page
connector, 1-11 · harmonic, 6-115 · monitor connector, 1-10 · monitors, 11-6 · multiple-controller capability, 11-17 · operation concepts, 6-1 · multiple measurements · plotted, 4-26 · operations · printed, 4-7 · instrument, 6-109{116 · optimizing · calibration standards, 5-6 · measurement results, 5-1 · naming conventions, 11-21 · measurements, 4-43 · option · noise oor · reducing, 5-56 · service, 1-13 · noise reduction, 6-44 · averaging, 6-44 · IF bandwidth, 6-45 · smoothing, 6-45 · non-coaxial measurements, 2-76 · non-insertable devices · calibrating for, 6-101 · noninsertable devices · calibrating for, 5-41 · non-volatile memory, 12-2 · available, 12-2 · number of points · option 001 · speci cations, 7-34 · option 002, 6-115 · o set, 6-7 · Option 010, 6-117{137 · limits, 6-108 · R-Channel, 3-4 · options available, 1-12 · specify, 6-80 · output · video, 7-35 · output power · speci cations, 7-27 · one-port calibration, 6-73 · outputting · measurement results, 4-1 · measurements connections, 3-9 · printer, 4-25 · plot les, 4-20 · on-wafer measurements, 2-76 · operating conditions, 7-36 · outputting plot les · operating parameters · operation · overview · Index-16
Page 669 of the Agilent Technologies 8753D user manual — PEN NUM TEXT , 4-13
Page 669 · click to enlarge
Page 669

P

P1 and P2 on the plotter, 4-15

con gurations, 4-4, 4-8, 11-8

connections to analyzer, 4-8

page of values printed or plotted, 4-30

parallel poll non response (PPO), 11-18

phase and magnitude response measurement,

GPIO mode, 6-105, 6-139

phase distortion and electrical length

de ned boundaries, 4-15

defaults for plotting, 4-16

defaults for printing, 4-6

operating:printing or plotting, 4-30

multiple measurements per page, 4-26

parameters set with markers, 2-16

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOT DATA , 4-12

how to view on a PC, 4-20

outputting from a PC to a plotter, 4-22

outputting from a PC to a printer, 4-23

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PLOT GRAT , 4-12

NNNNNNNNNNNNNNNNNNNNNNNNNN

PLOT MEM , 4-12

NNNNNNNNNNNNNNNNNNNNNNNNNN

PLOT MKR , 4-12

pass control capabilities (C10), 11-18

con guration, 4-8, 11-9

indicators in limit testing, 2-44

pen number settings, 4-13

pass fail indicators on display, 1-9

plotter P1 and P2, 4-15

PC? status notation, 1-8

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PC status notation, 1-8

PLOT TEXT , 4-12

P# status notation, 1-8

pen number settings, 4-13

a measurement to disk, 4-19

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PEN NUM DATA , 4-13

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

components de ned, 4-12

PEN NUM GRATICULE , 4-13

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PEN NUM MARKER , 4-13

multiple measurements on a full page,

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PEN NUM MEMORY , 4-13

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

multiple measurements per page, 4-17

PEN NUM TEXT , 4-13

con guring to plot, 4-10, 11-11

to an HPGL compatible printer, 4-18

plotting multiple measurements per page,

plotting, printing, and saving, 4-1

point limit creation, 2-41

Show 56 more labels / short text on this page
system, 6-2{3 · measurement accuracy, 5-3 · peripheral · addresses, 11-20 · page halves, 4-7 · peripheral equipment, 6-3 · page quadrants, 4-28 · peripherals, 11-1 · panel · connecting, 11-7 · rear, 1-10 · phase · parallel interface, 1-10 · speci cations, 7-29 · parallel port, 7-35 · copy mode, 6-105 · phase characteristics, 7-31 · plotting, 6-105 · measurement, 2-30 · printing, 6-105 · phase format, 6-30 · TTL input, 6-140 · physical dimensions, 7-37 · TTL output, 6-140 · plot · use of, 6-105 · aborting, 4-30 · parameters · de nition, 4-12 · lename convention, 4-26 · solving problems, 4-32 · center frequency, 2-18 · CW frequency, 2-21 · plot les · display reference, 2-20 · electrical delay, 2-21 · frequency span, 2-19 · start frequency, 2-17 · stop frequency, 2-17 · pass control · mode, 6-104 · plot speed, 4-15 · pass-control mode, 11-20 · plotter · pass/fail · display location, 1-9 · line types, 4-14 · plotters, 11-4 · plotting · arrays, 4-12 · measurement results, 4-3 · page quadrants, 4-28 · pen plotter · start, 4-16 · performance veri cation · Index-17
Page 670 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 670 · click to enlarge
Page 670

PPO (does not respond to parallel poll, 11-18

data:how to reduce, 5-52

pre-raw data arrays, 6-6

LIN MKR, 2-15

LOG MKR, 2-15

PRESET

menu map, 8-12

principles of microwave connector care, 2-2

polar or Smith format markers, 2-12

port 1 and port2, 1-5

multiple measurements per page, 4-7

NNNNNNNNNNNNNNNNNNNNNNNNNNNNN

PRINT ALL use, 4-31

increasing test port, 5-56

con guration, 4-4, 4-9, 11-8

outputting single page plots to, 4-24

using to output multiple plots to a single

power cord receptacle with fuse, 1-10

power level, calibrated, 6-95

printer connections to analyzer, 4-3

calibration for mixer measurements, 3-6

power meter calibration, 6-95

printing a sequence, 2-60

compatible sweep types, 6-95

printing or plotting list values or operating

continuous correction mode, 5-37

continuous sample mode, 6-96

printing, plotting, and saving, 4-1

print/plot time stamp, 4-30, 11-13

PRm status notation, 1-8

loss of calibration data, 5-33

modes of operation, 6-96

probe power source connector location, 1-5

power loss correction list, 6-98

with disk storage, 4-43

power sensor calibration data, 6-98

with printing or plotting, 4-32

receiver to measure absolute power, 5-38

sample-and-sweep correction mode, 5-36

aborting a print or plot process, 4-30

activating a xed marker, 2-11

activating display markers, 2-10

basic measurement sequence, 2-3

power (output) characteristics, 7-27

calculating data statistics, 2-26

calibrating the analyzer receiver to measure

changing the sequence title, 2-58

clearing a sequencing from memory, 2-57

compensating for directional coupler

calibration factor list, 6-98

con guring a plot function, 4-8

con guring a print function, 4-3

Show 61 more labels / short text on this page
points · precautions · pola format · electrostatic, 7-36 · markers, 2-14 · polar chart markers · preset · key location, 1-5 · · Re/Im MKR, 2-15 · preset state, 12-6 · polar format, 6-32 · print · aborting, 4-30 · port coupling, 6-15 · de nition, 4-5 · port extensions, 5-3 · port power · solving problems, 4-32 · increasing, 5-56 · power · printer · coupling, 6-15 · color, 4-6 · line, 7-36 · mode, 4-6 · menu, 6-13 · probe, 7-34 · page, 4-25 · power meter · printers, 11-4 · calibration, 5-33 · printing · default setting, 4-6 · con guration, 11-12 · measurement results, 4-3 · start, 4-6 · accuracy, 6-98, 6-99 · two measurements, 4-7 · applications, 6-95 · parameters, 4-30 · interpolation, 6-96 · loss of, 6-96 · probe power, 7-34 · power level, 6-95 · problems · procedure · sample-and-sweep mode, 6-97 · speed, 6-98 · stimulus parameters, 6-98 · power ranges, 6-13 · auto, 6-13 · absolute power, 5-38 · manual, 6-13 · power sensor · calibration data, 5-34 · response, 5-35 · power sweep, 6-23 · power sweep range · de ning, 6-11 · Index-18
Page 671 of the Agilent Technologies 8753D user manual — Index
Page 671 · click to enlarge
Page 671

constructing a loop structure in a sequence,

modifying TRMM calibration standards,

coupling and uncoupling display markers,

o setting limit lines, 2-44

outputting a single page of values, 4-30

creating a sequence, 2-54

outputting multiple plots to a single page

creating at limit lines, 2-37

using a printer, 4-25

creating single point limits, 2-41

outputting plot les, 4-20

creating sloping limit lines, 2-39

outputting plot les from a PC to a plotter,

de ning line types, 4-14

de ning the plot, 4-12

outputting plot les from a PC to a printer,

de ning the print, 4-5

deleting an instrument state, 4-41

outputting single page plots using a printer,

deleting a single instrument state le, 4-41

deleting limiting segments, 2-42

outputting the entire list of values, 4-31

deviation from linear phase measurement,

plotting a measurement to disk, 4-19

plotting measurements in page quadrants,

dividing measurement data by the memory

plotting multiple measurements on a full

editing a sequence, 2-56

editing limit segments, 2-42

plotting multiple measurements per page,

entering the power sensor calibration data,

plotting to an HPGL compatible printer,

error-correction for full two-port

printing a sequence, 2-60

error-correction for one-port re ection

printing multiple measurements per page,

error-correction for TRL*, 5-23

printing one measurement per page, 4-6,

error-correction for TRM*, 5-23

formatting a disk, 4-43

printing or plotting list values or operating

frequency response error-corrections, 5-8

gating in time domain, 2-70

purging a sequence from disk, 2-60

inserting a sequence command, 2-56

ratioing measurements in channel 1 and

limit testing using a sequence, 2-64

loading a sequence from disk, 2-60

recalling a le, 4-42

making accurate measurements of

renaming an instrument state le, 4-42

electrically long devices, 5-48

re-saving an instrument state, 4-41

measuring devices using the time domain,

resetting plotting parameters to default

measuring electrical length and phase

resetting the printing parameters to default

measuring gain compression, 2-45

response and isolation error-correction for

measuring group delay, 2-33

re ection measurements, 5-13

measuring magnitude and insertion phase

response and isolation error-correction for

measuring re ection response in time

response error-correction for re ection

measuring reverse isolation and gain, 2-49

response error-correction for transmission

measuring swept harmonics, 2-66

measuring transmission response in time

reviewing limit line segments, 2-43

running a limit test, 2-43

mixer group delay or phase, 3-24

running a sequence, 2-55

modifying a sequence command, 2-57

saving a data trace to the display memory,

modifying calibration standards, 5-26

modifying TRL calibration standards, 5-28

saving an instrument state, 4-35

searching for a target amplitude, 2-24

Show 17 more labels / short text on this page
trace, 2-7 · page, 4-26 · 4-17, 4-26 · measurements, 5-20 · measurements, 5-17 · parameters, 4-30 · 2, 2-7 · values, 4-16 · distortion, 2-30 · values, 4-6 · response, 2-27 · transmission measurements, 5-15 · domain, 2-73 · measurements, 5-8 · measurements, 5-10 · domain, 2-68 · Index-19
Page 672 of the Agilent Technologies 8753D user manual — NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Page 672 · click to enlarge
Page 672

searching for bandwidth, 2-25

programming speci cations, 7-33

searching for maximum amplitude, 2-22

searching for minimum amplitude, 2-23

P? status notation, 1-8

setting center frequency with markers,

purging a sequence from disk, 2-60

purpose and use of di erent error-correction

setting CW frequency using markers, 2-21

setting display reference using markers,

Q

setting electrical delay using markers, 2-21

setting frequency span with markers, 2-19

quarter page plots, 4-28

setting measurement parameters, 2-3

R

setting start frequency with markers, 2-17

setting stop frequency with markers, 2-17

rack mount cabinet, 11-4

setting the frequency range, 2-3

rack mount ange kit

setting the measurement type, 2-4

with handles option, 1-12

setting the source power, 2-4

without handles option, 1-12

setting up a color printer, 4-6

solving problems with disk storage, 4-43

dynamic:how to increasing, 5-56

stopping a sequence, 2-55

storing a sequence on a disk, 2-59

subtracting memory trace from

ratio measurement in channel 1 and 2, 2-7

measurement data trace, 2-7

raw data arrays, 6-6

R channel and source calibration connections,

test with limit lines, 2-36

titling the active channel display, 2-8

R channel connector locations, 1-5

titling the displayed measurement, 4-29

R-Channel inputs , di erence between internal

using analyzer display markers, 2-9

R-Channel power o set, 3-4

using continuous and discrete markers,

using delta () markers, 2-11

features and connectors, 1-10

NNNNNNNNNNNNNNNNNNNNNNNNNN

MKR ZERO to activate a xed

rear panel connectors, 7-34

recalling a le, 4-42

using polar format markers, 2-14

using the tuned receiver mode, 2-51

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REF=FIXED MKR key to activate

a xed reference marker, 2-12

receiver calibration procedure, 5-38

viewing both measurement channels, 2-5

receiver con guration option, 1-12

viewing plot les on a PC, 4-20

viewing the measurement data and memory

LO to RF isolation for mixers, 3-33

mixer conversion compression, 3-28

number of measurement points, 5-52

mixer RF feedthrough, 3-35

receiver noise oor, 5-56

power meter calibration for mixer

programming-command to key-function cross

Show 42 more labels / short text on this page
programs · example, 11-6 · sequencing, 3-17 · procedures, 5-5 · quadrants plotted, 4-28 · range · time domain, 6-130 · ratio calculations, 6-5 · test sequencing, 2-53 · tracking amplitude, 2-25 · and external, 3-4 · real format, 6-34 · rear panel · using · reference marker, 2-13 · receiver, 6-3 · receiver calibration, 5-11 · using the · connections, 3-14 · receiver crosstalk · reducing, 5-57 · trace, 2-6 · receiver noise oor · procedures · reducing, 5-56 · error-correcting measurements, 5-7 · reduce · averaging factor, 5-52 · reducing · measurements, 3-6 · receiver crosstalk, 5-57 · trace noise, 5-57 · processing details, 6-5 · reducing noise, 6-44 · programming, 11-14 · averaging, 6-44 · IF bandwidth, 6-45 · reference, 9-48 · smoothing, 6-45 · programming interface, 11-14 · reference · Index-20
Page 673 of the Agilent Technologies 8753D user manual — SCALE REF
Page 673 · click to enlarge
Page 673

(10 MHz) adjust, 1-11

return to HP calibration agreement option,

(10 MHz) output, 1-11

level of display, 1-8

return to HP calibration option, 1-13

return to HP full service agreement option,

return to HP repair agreement option, 1-13

reverse isolation and gain measurement,

using bandpass mode, 6-119

reviewing the limit line segments, 2-43

using low pass mode, 6-123

RF

re ection response in time domain, 2-73

how de ned for mixers, 3-2

uncorrected speci cations, 7-2

RL1 (complete remote/local capability), 11-18

R (remote operation), 11-19

RS-232 (serial) interface, 1-10

Re/Im MKR, 2-15, 2-16

rules for code naming, 11-21

a limit test, 2-43

S

relative velocity for time domain, 2-75

sample-and-sweep correction mode, 5-36

remote enable (REN) control line, 11-17

remote/local capability (RL1), 11-18

for power meter calibration, 5-36

remote operation (R), 11-19

remote programming speci cations, 7-33

save a data trace to the display memory, 2-6

renaming a le, 4-42

SA VE/RECALL

menu map, 8-11

REN (remote enable) control line, 11-17

a user kit, 5-28

places where you can, 4-33

to a computer, 4-34

re-save a le, 4-41

user kit to disk, 5-28

saving measurement results including

plotting parameters to default values, 4-16

saving, printing, and plotting, 4-1

scale/div . area of display, 1-8

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

SCALE PLOT , 4-15

calibration connections, 3-18, 3-34, 3-36

scale reference menu, 6-38

error-correction for re ection

SCALE REF

menu map, 8-12

searching for values with markers, 2-22

function block location, 1-4

response and isolation calibration, 6-73

second portion of conversion compression

restarting a calibration, 6-75

of memory data, 4-33

Show 56 more labels / short text on this page
markers, 2-11 · reference documents, 6-163 · reference plane · extending, 5-3 · re ection measurements · re ection tracking · RF feedthrough, 3-35 · R+jX MKR, 2-16 · register · contents, 4-33 · RPG knob, 6-10 · register data · retention, 4-33 · relative · running · marker mode, 2-11 · relative velocity factor · a sequence, 2-55 · adjusting of, 6-119 · remote control, 11-14 · sample-and-sweep mode · sampler/IF correction, 6-5 · · saving · repeatability · a le, 4-33 · connector, 5-2 · repetitive attenuator switching · allowing, 6-19 · reset · calibration, 4-36 · printing parameters, 4-6 · resolution · scale, 6-7 · time domain, 6-131 · response · · measurements, 5-8 · error-correction for transmission · bandwidth, 2-25 · measurements, 5-10 · maximum amplitude, 2-22 · minimum amplitude, 2-23 · target amplitude, 2-24 · response calibration, 6-73 · tracking amplitude, 2-25 · response function, 6-25{46 · measurement connections, 3-30 · results of measurement · segment · saving, 4-36 · deleting, 2-42 · retention · segment menu, 6-22 · selecting · Index-21
Page 674 of the Agilent Technologies 8753D user manual — LOG MKR, 2-16
Page 674 · click to enlarge
Page 674

selecting line types, 4-14

shortened sweep time, 5-50

sensor calibration data, 5-34

signal ow in a mixer example, 3-33

SEQ

menu map, 8-13

single bus concept, 11-19

single page of values printed or plotted, 4-30

single point limits, 2-41

six gigahertz operation option, 1-12

sloping limit lines, 2-39

sequence of measurement, 2-3

Smith chart markers, 2-15

cascading multiple sequences, 2-61

changing the sequence title, 2-58

LIN MKR, 2-16

clearing a sequence from memory, 2-57

LOG MKR, 2-16

decision making functions, 6-142

Smith or polar format markers, 2-12

Smo status notation, 1-8

editing a sequence, 2-56

gosub sequence command, 6-140

labels of display, 1-9

inserting a command, 2-56

limit test sequence, 2-64

softkey de nitions, 9-2

loading a sequence from disk, 2-60

loop counter sequence, 2-62

softkey menu maps, 8-1

softkeys and corresponding front panel access

modifying a command, 2-57

printing a sequence, 2-60

purging a sequence from disk, 2-60

reading using HP-IB, 6-144

running a test sequence, 2-55

for printing or plotting problems, 4-32

special functions menu, 6-142

stopping a sequencing, 2-55

disk storage problems, 4-43

storing a sequence on a disk, 2-59

TTL input for decision making, 6-140

uncorrected speci cations, 7-2

TTL I/O menu, 6-140

TTL out menu, 6-142

frequency:decrease, 5-50

with loop counter, 6-139

span setting for gating, 2-70

serial number plate location, 1-10

serial (RS-232) interface, 1-10

service and support options, 1-13

3.5 mm test ports, 7-8

service request asserted by the analyzer (S),

50 type-N test ports, 7-5

75 type-F test ports, 7-14

service request (SRQ) control line, 11-17

75 type-N test ports, 7-11

setting gate parameters, 2-70

7 mm test port, 7-2

setting HP-IB addresses, 11-20

setting measurement parameters using

external source mode, 7-29

sex of type-N calibration standard, 5-6

SH1 (full-source handshake), 11-18

frequency o set, 7-29

shape of gate, 2-70

Show 46 more labels / short text on this page
calibration kits, 6-75 · shipment weight, 7-37 · · sequence · creation, 2-54 · size of, 6-139 · using HP-IB, 6-144 · smith chart · sequencing · format, 6-31 · autostarting, 6-139 · G+jB MKR, 2-16 · command information, 6-138 · Re/Im MKR, 2-16 · concepts, 6-138{144 · R+jX MKR, 2-16 · deleting commands, 2-56 · smoothing, 6-7, 6-45 · editing, 6-139 · softkey · label location, 1-9 · HP-GL commands, 6-143 · location, 1-4 · softkey locations, 9-69 · menu, 6-140 · key, 9-69 · software · sample, 11-6 · solutions · solving · source, 6-2 · subprograms, 2-61 · source attenuator · test procedure, 2-53 · switch protection, 6-19 · titles, 6-139 · source match, 6-52 · span · S-parameters, 6-26 · speci cations, 7-1 · dynamic range, 7-1 · error-corrected, 7-2 · markers, 2-16 · frequency characteristics, 7-27 · group delay, 7-31 · Index-22
Page 675 of the Agilent Technologies 8753D user manual — HP-IB, 6-104
Page 675 · click to enlarge
Page 675

harmonics (option 002), 7-28

HP-IB, 6-104

* status notation, 1-8

steps of making a measurement, 2-3

spectral purity characteristics, 7-27

type-N test ports, 7-4

uncorrected test port, 7-2

setting with markers, 2-17

o set menus, 6-80

spectral purity characteristics, 7-27

a print or plot, 4-30

spectrum of RF ,LO and IF signals present in

a conversion loss measurement, 3-7

stop values possible, 1-7

SR1 (complete service request capabilities),

a sequence on a disk, 2-59

SRQ (service request) control line, 11-17

to external disk, 11-5

S (service request asserted by the analyzer),

to internal disk, 11-5

to internal memory, 11-5

full two port error-correction, 5-20

one port re ection error-correction, 5-18

structure of HP-IB bus, 11-16

subtract memory trace from the measurement

response and isolation error-correction

for& re ection measurements, 5-14

support and service options, 1-13

response and isolation error-correction

for& transmission measurements, 5-16

de nition of, 6-5

response error-correction for re ection

how to set auto sweep time, 5-51

how to set chop sweep, 5-54

response error-correction for transmission

de ning menus, 6-78

standard system maintenance service option,

start frequency set with markers, 2-17

how to set, 5-53

start values possible, 1-6

Show 68 more labels / short text on this page
statistics calculation, 2-26 · impedance, 7-28 · status indicators, 11-19 · magnitude, 7-29 · measurement port, 7-2 · noise level, 7-28 · status notations, 1-7 · option 001, 7-34 · step attenuator, 6-2 · output power, 7-27 · switch protection, 6-19 · phase, 7-29, 7-31 · step keys, 6-10 · remote programming, 7-33 · stimulus · spurious signals, 7-27 · function block, 1-4 · state, 5-5 · uncorrected, 7-2 · stimulus function, 6-11{24 · stimulus menu, 6-12 · specify · stop frequency · class menu, 6-81 · stopping · a sequencing, 2-55 · speed increased, 5-50 · storage · split display, 2-5 · locations, 4-33 · spurious signals · mediums, 4-33 · speci cations, 7-27 · storage conditions, 7-36 · storing · standard connections for · storing data · methods of, 12-1 · receiver calibration, 5-11 · data trace, 2-7 · sweep · measurement, 5-9 · sweep modes · measurements, 5-10 · alternate, 6-23, 6-100 · standard de nitions · chop, 6-23, 6-100 · entering, 5-26 · sweep speed · examining, 5-26 · increase, 5-50 · modifying, 5-26 · increasing, 5-50 · standards · sweep time · calibration, 5-6 · auto, 6-16 · manual, 6-16 · minimum, 6-16 · sweep type · starting · sweep types, 6-21 · a plot, 4-16 · CW time, 6-23 · a print, 4-6 · linear, 6-21 · list frequency, 6-22 · Index-23
Page 676 of the Agilent Technologies 8753D user manual — T
Page 676 · click to enlarge
Page 676

swept harmonics measurement, 2-66

swept IF conversion loss measurement

3.5 mm test port speci cations, 7-8

50 type-N test port speci cations, 7-5

swept power conversion compression

75 type-F test port speci cations, 7-14

75 type-N test port speci cations, 7-11

swept RF/IF mixer measurement, 3-7

test sequence connector location, 1-11

syntax for commands, 11-21

test sequence output (TEST SEQ), 7-35

cascading multiple sequences, 2-61

changing the sequence title, 2-58

clearing a sequence from memory, 2-57

decision making functions, 6-142

editing a sequence, 2-56

gosub sequence command, 6-140

how to change, 5-56

GPIO, 6-139

how to widen, 5-52

system controller capabilities (C1,C2,C3),

inserting a command, 2-56

limit test sequence, 2-64

system-controller mode, 11-19, 11-20

loading a sequence from disk, 2-60

SYSTEM

loop counter sequence, 2-62

SYSTEM

menu map, 8-15

modifying a command, 2-57

T

printing a sequence, 2-60

T6 (basic talker), 11-18

talker interface function, 11-15

purging a sequence from disk, 2-60

reading using HP-IB, 6-144

running a sequence, 2-55

special functions menu, 6-142

talk mode (T), 11-19

stopping a sequence, 2-55

target amplitude search, 2-24

storing a sequence on a disk, 2-59

TE0 (no extended talker capabilities), 11-18

TTL I/O menu, 6-140

TTL out menu, 6-142

with loop counter, 6-139

testing with limit lines, 2-36

test set switch, controlling the, 5-54

test using limits, 2-43

speci cations:50 , 7-2

speci cations:7 mm, 7-2

test port bias input (BIAS CONNECT), 7-35

test port coupling, 6-15

bandpass mode:re ection, 6-119

test port dimensions, 7-33

bandpass mode:transmission, 6-121

Show 57 more labels / short text on this page
logarithmic, 6-22 · test port power · power, 6-23 · increasing, 5-56 · test ports · example, 3-16 · measurement example, 3-32 · test sequence · switch · size of, 6-139 · protection, 6-19 · using HP-IB, 6-144 · SWR format, 6-33 · synthesized source, 6-2 · test sequencing · system · autostarting, 6-139 · cabinet, 11-4 · characterization, 5-7 · controller mode, 6-104 · menu, 6-106 · command information, 6-138 · overview, 6-2 {3 · concepts, 6-138{144 · testmobile, 11-4 · system accessories, 11-4 · deleting commands, 2-56 · systematic errors, 5-4 · editing, 6-139 · characterizing, 6-53 · system bandwidth · HP-GL commands, 6-143 · · key, 6-109 · · menu, 6-140 · procedures, 2-53 · talker/listener · mode, 6-104 · talker/listener mode, 11-20 · subprograms, 2-61 · techniques · titles, 6-139 · optimizing measurements, 4-43 · temperature conditions, 7-36 · temperature drift, 5-2 · test xtures · test set, 6-3 · transistors, 11-3 · hold mode, 6-19 · interconnect location, 1-11 · testmobile, 11-4 · test port · time domain · bandpass, 6-119 · bandpass mode, 6-117 · Index-24
Page 677 of the Agilent Technologies 8753D user manual — TTL I/O, 6-140
Page 677 · click to enlarge
Page 677

forward transform mode, 6-118

time domain bandpass, 6-117

time domain low pass impulse, 6-117

time domain low pass step, 6-117

transistor test xtures, 11-3

low pass impulse mode, 6-117

low pass mode , fault location

using bandpass mode, 6-121

using low pass mode, 6-125

low pass mode , re ection, 6-123

transmission response in time domain, 2-68

low pass mode , transmission, 6-125

low pass , setting frequency range, 6-122

uncorrected speci cations, 7-2

low pass step mode, 6-117

tri-state drivers (E2), 11-18

re ection response, 2-73

relative velocity factor, 6-119

fabricating and de ning standards, 6-91

how it works, 6-86

improving raw load match, 6-88

transmission response measurement, 2-68

improving raw source match, 6-88

time for measurements, 7-33

reason for using, 6-85

time stamp, 4-30, 11-13

requirements for standards, 6-90

area of display, 1-8

TRL*/LRM* two-port calibration, 6-74

TRL standard modi cation, 5-28

the active channel display, 2-8

TRM standard modi cation, 5-30

the displayed measurement, 4-29

to use fast 2-port calibration, 5-54

solving for printing or plotting, 4-32

with disk storage, 4-43

tsH status notation, 1-8

divide measurement data by the memory

T (talk mode), 11-19

TTL I/O, 6-140

trace math operation, 6-6

tuned receiver mode, 3-17, 6-111

compatible sweep types, 2-51

external source requirements, 2-52

transfer of data, 11-16

type-F (75 ) test ports

measurement uncertainties, 7-14, 7-17,

CW time-to-frequency domain, 6-134

type-N (50 ) test ports

Show 62 more labels / short text on this page
concepts, 6-117{137 · transform menu, 6-117 · transform modes · gating, 6-133 · gating procedure, 2-70 · general theory, 6-118 · low pass, 6-122 · transmission measurements · measurements, 6-123 · transmission tracking · trigger menu, 6-18 · masking, 6-127 · measurements, 2-68 · TRL calibration, 2-76 · option, 1-12 · TRL* error-correction, 5-23 · range, 6-130 · TRL*/LRM* calibration, 6-85 · error model, 6-86 · resolution, 6-131 · transform, 6-6 · velocity factor, 2-75 · isolation, 6-87 · windowing, 6-128 · load match, 6-88 · time domain range · options, 6-93 · de ning, 6-11 · procedure, 6-90 · time-to-frequency domain, 6-118 · source match, 6-88 · title · terminology, 6-85 · of sequence, 2-58 · TRM* error-correction, 5-24 · titling · trouble · trace · dual, 6-8 · tsH annotation, 6-19 · trace math, 2-7 · trace, 2-7 · trace noise · reducing, 5-57 · description, 2-51 · trace viewing, 2-6 · tracking, 6-53 · making measurements, 2-51 · an amplitude, 2-25 · mixer testing, 6-149 · transfer formats, 7-33 · test setup, 2-51 · transform, 6-6 · 7-20, 7-23 · frequency-to-time domain, 6-117 · type-F test ports · time-to-frequency domain, 6-118 · speci cations, 7-14 · transform measurements · forward, 6-135 · speci cations, 7-5 · Index-25
Page 678 of the Agilent Technologies 8753D user manual — W
Page 678 · click to enlarge
Page 678

type-N (75 ) test ports

type-N calibration standard sex, 5-6

type-N test ports (50 )

how to set, 5-53

typical calibration kit standard and

velocity factor for time domain

U

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

REF=FIXED MKR , 2-12

veri cation kit, 11-2

verifying performance of calibration, 6-84

3.5 mm test port measurements, 7-8

video output (EXT MON), 7-35

50 type-N test port measurements, 7-5

75 type-F test ports, 7-14, 7-17, 7-20,

both channels with a split display, 2-5

both measurement channels, 2-5

75 type-N test ports, 7-11

plot les on a PC, 4-20

7 mm test port measurements, 7-2

the measurement data and memory trace,

voltage selector switch, 1-10

" status notation, 1-8

W

up converter port connections, 3-3

widen system bandwidth, 5-52

Show 31 more labels / short text on this page
saving, 5-28 · measurement uncertainties, 7-11 · user kit saved · speci cations, 7-11 · to disk, 5-28 · using · type-N test ports · AmiPro, 4-21 · speci cations, 7-4 · Freelance, 4-22 · measurement uncertainties, 7-5 · type of sweep · valid characters, 11-22 · values · list, 4-30 · corresponding number, 5-27 · vector error-correction, 6-6 · measurements, 2-75 · uncertainties · viewing · understanding S-parameters, 6-26 · units, 11-23 · volatile memory, 12-1 · units terminator, 6-9 · user kit · weight, 7-37 · modi ed, 5-28 · modifying, 5-28 · windowing · time domain, 6-128 · Index-26

Ask about the Agilent Technologies 8753D

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the Agilent Technologies 8753D

Be the first to ask — other owners or our team will answer.

Original document

Agilent Technologies 8753D manual PDF

678 pages·4.8 MB·Network Router·PDF
Click any page above to open it here
Page 1 / …
100%
Loading pages…