Hewlett-Packard to Agilent Technologies Transition
This documentation supports a product that previously shipped under the Hewlett- Packard company brand name. The brand name has now been changed to Agilent Technologies. The two products are functionally identical, only our name has changed. The document still includes references to Hewlett-Packard products, some of which have been transitioned to Agilent Technologies.


HP 8753D Network Analyzer
HP P art No . 08753-90257 Supersedes October 1997
Printed in US A December 1997
User's Guide

The information contained in this document is subject to change without notice .
Hewlett-P ackard makes no warranty of any kind with regard to this material, including
but not limited to , the implied warranties of merchantability and tness for a particular
purpose . Hewlett-P ackard shall not be liable for errors contained herein or for incidental or
consequential damages in connection with the furnishing, performance , or use of this material.
c Copyright Hewlett-P ackard Company 1994, 1995, 1997
All Rights Reserved. Reproduction, adaptation, or translation without prior written permission
is prohibited, except as allowed under the copyright laws .
1400 F ountaingrove P arkway , Santa Rosa, CA 95403-1799, USA

Hewlett-P ackard Company certi es that this product met its published speci cations at the
time of shipment from the factory . Hewlett-P ackard further certi es that its calibration
measurements are traceable to the United States National Institute of Standards and
T echnology , to the extent allowed by the Institute's calibration facility , and to the calibration
facilities of other International Standards Organization members .
This Hewlett-P ackard instrument product is warranted against defects in material and
workmanship for a period of one year from date of shipment. During the warranty period,
Hewlett-P ackard Company will, at its option, either repair or replace products which prove to
be defective .
F or warranty service or repair , this product must be returned to a service facility designated by
Hewlett-P ackard. Buyer shall prepay shipping charges to Hewlett-P ackard and Hewlett-P ackard
shall pay shipping charges to return the product to Buyer . However , Buyer shall pay all
shipping charges , duties , and taxes for products returned to Hewlett-P ackard from another
country .
Hewlett-P ackard warrants that its software and rmware designated by Hewlett-P ackard for
use with an instrument will execute its programming instructions when properly installed on
that instrument. Hewlett-P ackard does not warrant that the operation of the instrument, or
software , or rmware will be uninterrupted or error-free .
Limit a tion of W arranty
The foregoing warranty shall not apply to defects resulting from improper or inadequate
maintenance by Buyer , Buyer-supplied software or interfacing, unauthorized modi cation or
misuse , operation outside of the environmental speci cations for the product, or improper
site preparation or maintenance .
NO OTHER W ARRANTY IS EXPRESSED OR IMPLIED . HEWLETT-P A CKARD SPECIFICALL Y
DISCLAIMS THE IMPLIED W ARRANTIES OF MERCHANT ABILITY AND FITNESS FOR A
P ARTICULAR PURPOSE.
THE REMEDIES PRO VIDED HEREIN ARE BUYER'S SOLE AND EXCLUSIVE REMEDIES.
HEWLETT-P A CKARD SHALL NOT BE LIABLE FOR ANY DIRECT , INDIRECT , SPECIAL,
INCIDENT AL, OR CONSEQUENTIAL D AMA GES, WHETHER B ASED ON CONTRA CT , TORT ,
OR ANY OTHER LEGAL THEORY .
Certi cation · W arranty · Ex clusive Remedies

Clean the cabinet, using a damp cloth only .
Product maintenance agreements and other customer assistance agreements are available for
Hewlett-P ackard products .
F or any assistance , contact your nearest Hewlett-P ackard Sales and Service Oce .
Maintenance · Assistance

By internet, phone, or fax, get assistance with all your test and measurement needs.
Online assistance: www.agilent.com/find/assist
United States (tel) 1 800 452 4844
Latin America (tel) (305) 269 7500 (fax) (305) 269 7599
Canada (tel) 1 877 894 4414 (fax) (905) 282-6495
Europe (tel) (+31) 20 547 2323 (fax) (+31) 20 547 2390
New Zealand (tel) 0 800 738 378 (fax) (+64) 4 495 8950
Japan (tel) (+81) 426 56 7832 (fax) (+81) 426 56 7840
Australia (tel) 1 800 629 485 (fax) (+61) 3 9210 5947
Asia Call Center Numbers
Country Phone Number Fax Number
Singapore 1-800-375-8100 (65) 836-0252
Malaysia 1-800-828-848 1-800-801664
Philippines (632) 8426802 1-800-16510170 (PLDT Subscriber Only)
(632) 8426809 1-800-16510288 (PLDT Subscriber Only)
Thailand (088) 226-008 (outside Bangkok) (662) 661-3999 (within Bangkok)
(66) 1-661-3714
Hong Kong 800-930-871 (852) 2506 9233
Taiwan 0800-047-866 (886) 2 25456723
People’s Republic of China
800-810-0189 (preferred) 10800-650-0021 10800-650-0121
India 1-600-11-2929 000-800-650-1101
2 Chapter1

The following safety symbols are used throughout this manual. F amiliarize yourself with each
of the symbols and its meaning before operating this instrument.
correctly performed or adhered to , would result in damage to or destruction
of the instrument. Do not proceed beyond a caution note until the indicated
conditions are fully understood and met.
W arning denotes a hazard. It calls attention to a procedure which, if not
correctly performed or adhered to , could result in injury or loss of life .
Do not proceed beyond a warning note until the indicated conditions are
fully understood and met.
L The instruction documentation symbol. The product is marked with this symbol when it
is necessary for the user to refer to the instructions in the documentation.
\CE" The CE mark is a registered trademark of the European Community . (If accompanied by
a year , it is when the design was proven.)
\ISM1-A" This is a symbol of an Industrial Scienti c and Medical Group 1 Class A product.
\CSA" The CSA mark is a registered trademark of the Canadian Standards Association.
Safety Symbols · W arning · Instrument Markings

This is a Safety Class I product (provided with a protective earthing
ground incorporated in the power cord). The mains plug shall only be
inserted in a socket outlet provided with a protective earth contact. Any
interruption of the protective conductor , inside or outside the instrument,
is likely to make the instrument dangerous . Intentional interruption is
prohibited.
No operator serviceable parts inside . Refer servicing to quali ed
personnel. T o prevent electrical shock, do not remove covers .
Before switching on this instrument, make sure that the line voltage selector
switch is set to the voltage of the power supply and the correct fuse is
installed.
The opening of covers or removal of parts is likely to expose dangerous
voltages . Disconnect the instrument from all voltage sources while it is
being opened.
The power cord is connected to internal capacitors that may remain live
for 10 seconds after disconnecting the plug from its power supply .
F or continued protection against re hazard replace line fuse only with
same type and rating (F 3A/250V). The use of other fuses or material is
prohibited.
If this instrument is used in a manner not speci ed by Hewlett-P ackard
Co ., the protection provided by the instrument may be impaired.
This instrument has been designed and tested in accordance with IEC
Publication 348, Safety Requirements for Electronics Measuring Apparatus , and
has been supplied in a safe condition. This instruction documentation contains
information and warnings which must be followed by the user to ensure safe
operation and to maintain the instrument in a safe condition.

Chapter 1, \HP 8753D Description and Options ," describes features , functions , and available
options .
Chapter 2, \Making Measurements ," contains step-by-step procedures for making
measurements or using particular functions .
Chapter 3, \Making Mixer Measurements ," contains step-by-step procedures for making
calibrated and error-corrected mixer measurements .
Chapter 4, \Printing, Plotting, and Saving Measurement Results ," contains instructions
for saving to disk or the analyzer internal memory , and printing and plotting displayed
measurements .
Chapter 5, \Optimizing Measurement Results ," describes techniques and functions for
achieving the best measurement results .
Chapter 6, \Application and Operation Concepts ," contains explanatory-style information
about many applications and analyzer operation.
Chapter 7, \Speci cations and Measurement Uncertainties ," de nes the performance
capabilities of the analyzer .
Chapter 8, \Menu Maps ," shows softkey menu relationships .
Chapter 9, \Key De nitions ," describes all the front panel keys , softkeys , and their
corresponding HP-IB commands .
Chapter 10, \Error Messages ," provides information for interpreting error messages .
Chapter 11, \Compatible P eripherals ," lists measurement and system accessories , and
other applicable equipment compatible with the analyzer . Procedures for con guring the
peripherals , and an HP-IB programming overview are also included.
Chapter 12, \Preset State and Memory Allocation," contains a discussion of memory
allocation, memory storage , instrument state de nitions , and preset conditions .
Appendix A, \The CITI le Data F ormat and Key W ord Reference ," contains information on
the CITI le data format as well as a list of CITI le keywords .
User's Guide Overview

Network Analyzer Documentation Set
The Installation and Quick Start Guide

familiarizes you with the network analyzer's
front and rear panels , electrical and
environmental operating requirements , as well
as procedures for installing, con guring, and
verifying the operation of the analyzer .
The User's Guide shows how to make
measurements , explains commonly-used

features , and tells you how to get the most
performance from your analyzer .
The Quick Reference Guide provides a
summary of selected user features .

The Programmer's Guide provides
programming information including an HP-IB
programming and command reference as well
as programming examples .
The System V eri cation and T est Guide
provides the system veri cation and
performance tests and the P erformance T est
Record for your analyzer .




T o Title the A ctive Channel Display 2-8
Using Analyzer Display Markers 2-9
T o Use Continuous and Discrete Markers 2-9
T o A ctivate Display Markers 2-10
T o Use Delta () Markers 2-11
T o A ctivate a Fixed Marker 2-11
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
REF=FIXED MKR Key to activate a Fixed Reference Marker . . 2-12
MKR ZERO Key to A ctivate a Fixed Reference Marker 2-13
T o Couple and Uncouple Display Markers 2-14
T o Use P olar F ormat Markers 2-14
T o Use Smith Chart Markers 2-15
T o Set Measurement P arameters Using Markers 2-16
Setting the Start Frequency 2-17
Setting the Stop Frequency 2-17
Setting the Center Frequency 2-18
Setting the Frequency Span 2-19
Setting the Display Reference V alue 2-20
Setting the Electrical Delay 2-21
Setting the CW Frequency 2-21
T o Search for a Speci c Amplitude 2-22
Searching for the Maximum Amplitude 2-22
Searching for the Minimum Amplitude 2-23
Searching for a T arget Amplitude 2-24
Searching for a Bandwidth 2-25
Tracking the Amplitude that Y ou are Searching 2-25
T o Calculate the Statistics of the Measurement Data 2-26
Measuring Magnitude and Insertion Phase Response 2-27
Measuring the Magnitude Response 2-27
Measuring Insertion Phase Response 2-28
Measuring Electrical Length and Phase Distortion 2-30
Measuring Electrical Length 2-30
Measuring Phase Distortion 2-32
Deviation From Linear Phase 2-32
Group Delay 2-33
T esting A Device with Limit Lines 2-36
Setting Up the Measurement P arameters 2-36
Creating Flat Limit Lines 2-37
Creating a Sloping Limit Line 2-39
Creating Single P oint Limits 2-41
Editing Limit Segments 2-42
Deleting Limit Segments 2-42
Running a Limit T est 2-43
Reviewing the Limit Line Segments 2-43
A ctivating the Limit T est 2-43
O setting Limit Lines 2-44
Measuring Gain Compression 2-45
Measuring Gain and Reverse Isolation Simultaneously 2-49
Measurements Using the Tuned Receiver Mode 2-51
Typical test setup 2-51
Tuned receiver mode in-depth description 2-51
Frequency Range 2-51
Compatible Sweep Types 2-51
External Source Requirements 2-52
T est Sequencing 2-53
Using the · Using the · Contents-2

Creating a Sequence 2-54
Running a Sequence 2-55
Stopping a Sequence 2-55
Editing a Sequence 2-56
Deleting Commands 2-56
Inserting a Command 2-56
Modifying a Command 2-57
Clearing a Sequence from Memory 2-57
Changing the Sequence Title 2-58
Naming Files Generated by a Sequence 2-58
Storing a Sequence on a Disk 2-59
Loading a Sequence from Disk 2-60
Purging a Sequence from Disk 2-60
Printing a Sequence 2-60
Cascading Multiple Example Sequences 2-61
Loop Counter Example Sequence 2-62
Generating Files in a Loop Counter Example Sequence 2-63
Limit T est Example Sequence 2-64
Measuring Swept Harmonics 2-66
Measuring a Device in the Time Domain (Option 010 Only) 2-68
Transmission Response in Time Domain 2-68
Re ection Response in Time Domain 2-73
Non-coaxial Measurements 2-76
Where to Look for More Information 3-1
Measurement Considerations 3-2
Minimizing Source and Load Mismatches 3-2
Reducing the E ect of Spurious Responses 3-2
Eliminating Unwanted Mixing and Leakage Signals 3-2
How RF and IF Are De ned 3-2
Frequency O set Mode Operation 3-4
Di erences Between Internal and External R-Channel Inputs 3-4
P ower Meter Calibration 3-6
Conversion Loss Using the Frequency O set Mode 3-7
High Dynamic Range Swept RF/IF Conversion Loss 3-12
Fixed IF Mixer Measurements 3-17
Tuned Receiver Mode 3-17
Sequence 1 Setup 3-17
Sequence 2 Setup 3-21
Phase or Group Delay Measurements 3-24
Amplitude and Phase Tracking 3-27
Conversion Compression Using the Frequency O set Mode 3-28
Isolation Example Measurements 3-33
LO to RF Isolation 3-33
RF F eedthrough 3-35
Contents-3

Where to Look for More Information 4-2
Printing or Plotting Y our Measurement Results 4-3
Con guring a Print Function 4-3
De ning a Print Function 4-5
If Y ou are Using a Color Printer 4-6
T o Reset the Printing P arameters to Default V alues 4-6
Printing One Measurement P er P age 4-6
Printing Multiple Measurements P er P age 4-7
Con guring a Plot Function 4-8
If Y ou are Plotting to an HPGL/2 Compatible Printer 4-8
If Y ou are Plotting to a P en Plotter 4-10
If Y ou are Plotting to a Disk Drive 4-11
De ning a Plot Function 4-12
Choosing Display Elements 4-12
Selecting A uto-F eed 4-12
Selecting P en Numbers and Colors 4-13
Selecting Line Types 4-14
Choosing Scale 4-15
Choosing Plot Speed 4-15
T o Reset the Plotting P arameters to Default V alues 4-16
Plotting One Measurement P er P age Using a P en Plotter 4-16
Plotting Multiple Measurements P er P age Using a P en Plotter 4-17
If Y ou are Plotting to an HPGL Compatible Printer 4-18
Plotting a Measurement to Disk 4-19
T o Output the Plot Files 4-20
T o View Plot Files on a PC 4-20
Using AmiPro 4-21
Using Freelance 4-22
Outputting Plot Files from a PC to a Plotter 4-22
Outputting Plot Files from a PC to an HPGL Compatible Printer 4-23
Outputting Single P age Plots Using a Printer 4-24
Outputting Multiple Plots to a Single P age Using a Printer 4-25
Plotting Multiple Measurements P er P age From Disk 4-26
T o Plot Multiple Measurements on a Full P age 4-26
T o Plot Measurements in P age Quadrants 4-28
Titling the Displayed Measurement 4-29
Con guring the Analyzer to Produce a Time Stamp 4-30
Aborting a Print or Plot Process 4-30
Printing or Plotting the List V alues or Operating P arameters 4-30
If Y ou want a Single P age of V alues 4-30
If Y ou W ant the Entire List of V alues 4-31
Solving Problems with Printing or Plotting 4-32
Saving and Recalling Instrument States 4-33
Places Where Y ou Can Save 4-33
What Y ou Can Save to the Analyzer's Internal Memory 4-33
What Y ou Can Save to a Floppy Disk 4-33
What Y ou Can Save to a Computer 4-34
Saving an Instrument State 4-35
Saving Measurement Results 4-36
Contents-4

ASCII Data F ormats 4-39
CITI le 4-39
S2P Data F ormat 4-39
Re-Saving an Instrument State 4-41
Deleting a File 4-41
T o Delete an Instrument State File 4-41
T o Delete all Files 4-41
Renaming a File 4-42
Recalling a File 4-42
F ormatting a Disk 4-43
Solving Problems with Saving or Recalling Files 4-43
If Y ou are Using an External Disk Drive 4-43
Where to Look for More Information 5-2
Increasing Measurement A ccuracy 5-2
Connector Repeatability 5-2
Interconnecting Cables 5-2
T emperature Drift 5-2
Frequency Drift 5-3
P erformance V eri cation 5-3
Reference Plane and P ort Extensions 5-3
Measurement Error-Correction 5-4
Conditions Where Error-Correction is Suggested 5-4
Types of Error-Correction 5-4
Error-Correction Stimulus State 5-5
Calibration Standards 5-6
Compensating for the Electrical Delay of Calibration Standards 5-6
Clarifying Type-N Connector Sex 5-6
When to Use Interpolated Error-Correction 5-6
Procedures for Error-Correcting Y our Measurements 5-7
Frequency Response Error-Corrections 5-8
Response Error-Correction for Re ection Measurements 5-8
Response Error-Correction for Transmission Measurements 5-10
Receiver Calibration 5-11
Frequency Response and Isolation Error-Corrections 5-13
Response and Isolation Error-Correction for Re ection Measurements 5-13
Response and Isolation Error-Correction for Transmission Measurements . . . 5-15
One-P ort Re ection Error-Correction 5-17
Full Two-P ort Error-Correction 5-20
TRL* and TRM* Error-Correction 5-23
TRL Error-Correction 5-23
TRM Error-Correction 5-24
Modifying Calibration Kit Standards 5-26
De nitions 5-26
Outline of Standard Modi cation 5-26
Modifying Standards 5-26
Modifying TRL Standards 5-28
Modifying TRM Standards 5-30
P ower Meter Measurement Calibration 5-33
Entering the P ower Sensor Calibration Data 5-34
Editing Frequency Segments 5-34
Deleting Frequency Segments 5-35
Compensating for Directional Coupler Response 5-35
Using Sample-and-Sweep Correction Mode 5-36
Contents-5

Using Continuous Correction Mode 5-37
T o Calibrate the Analyzer Receiver to Measure Absolute P ower 5-38
Matched A dapters 5-39
Modify the Cal Kit Thru De nition 5-40
Calibrating for Noninsertable Devices 5-41
A dapter Removal 5-42
P erform the 2-port Error Corrections 5-43
Remove the A dapter 5-44
V erify the Results 5-45
Example Program 5-47
Making A ccurate Measurements of Electrically Long Devices 5-48
The Cause of Measurement Problems 5-48
T o Improve Measurement Results 5-48
Decreasing the Sweep Rate 5-48
Decreasing the Time Delay 5-49
Increasing Sweep Speed 5-50
T o Decrease the Frequency Span 5-50
T o Set the A uto Sweep Time Mode 5-51
T o Widen the System Bandwidth 5-52
T o Reduce the A veraging F actor 5-52
T o Reduce the Number of Measurement P oints 5-52
T o Set the Sweep Type 5-53
T o View a Single Measurement Channel 5-53
T o A ctivate Chop Sweep Mode 5-54
T o Use External Calibration 5-54
T o Use F ast 2-P ort Calibration 5-54
Increasing Dynamic Range 5-56
T o Increase the T est P ort Input P ower 5-56
T o Reduce the Receiver Noise Floor 5-56
Changing System Bandwidth 5-56
Changing Measurement A veraging 5-56
Reducing Trace Noise 5-57
T o A ctivate A veraging 5-57
T o Change System Bandwidth 5-57
Reducing Receiver Crosstalk 5-57
Reducing Recall Time 5-58
Understanding Spur A voidance 5-59
Where to Look for More Information 6-1
HP 8753D System Operation 6-2
The Built-In Synthesized Source 6-2
The Source Step Attenuator 6-2
The Built-In T est Set 6-3
The Receiver Block 6-3
The Microprocessor 6-3
Required P eripheral Equipment 6-3
Data Processing 6-4
Processing Details 6-5
The ADC 6-5
IF Detection 6-5
Ratio Calculations 6-5
Sampler/IF Correction 6-5
Sweep-T o-Sweep A veraging 6-5
Pre-Raw Data Arrays 6-6
Contents-6

Raw Arrays 6-6
V ector Error-correction (A ccuracy Enhancement) 6-6
Trace Math Operation 6-6
Gating (Option 010 Only) 6-6
The Electrical Delay Block 6-6
Conversion 6-6
Transform (Option 010 Only) 6-6
F ormat 6-7
Smoothing 6-7
F ormat Arrays 6-7
O set and Scale 6-7
Display Memory 6-7
A ctive Channel Keys 6-8
Dual Channel 6-8
Uncoupling Stimulus V alues Between Channels 6-8
Coupled Markers 6-8
Entry Block Keys 6-9
Units T erminator 6-9
Knob 6-10
Step Keys 6-10
6-10
6-10
6-10
6-10
Stimulus Functions 6-11
De ning Ranges with Stimulus Keys 6-11
Stimulus Menu 6-12
The P ower Menu 6-13
Understanding the P ower Ranges 6-13
A utomatic mode 6-13
Manual mode 6-13
P ower Coupling Options 6-15
Channel coupling 6-15
T est port coupling 6-15
Sweep Time 6-16
Manual Sweep Time Mode 6-16
A uto Sweep Time Mode 6-16
Minimum Sweep Time 6-16
Trigger Menu 6-18
Source Attenuator Switch Protection 6-19
Allowing Repetitive Switching of the Attenuator 6-19
Channel Stimulus Coupling 6-20
Sweep Type Menu 6-21
Linear Frequency Sweep (Hz) 6-21
Logarithmic Frequency Sweep (Hz) 6-22
List Frequency Sweep (Hz) 6-22
Segment Menu 6-22
P ower Sweep (dBm) 6-23
CW Time Sweep (Seconds) 6-23
Selecting Sweep Modes 6-23
Modifying List Frequencies 6-23
Edit list menu 6-23
Edit subsweep menu 6-24
Response Functions 6-25
S-P arameters 6-26

Understanding S-P arameters 6-26
The S-P arameter Menu 6-27
Analog In Menu 6-27
Conversion Menu 6-27
Input P orts Menu 6-28
The F ormat Menu 6-29
Log Magnitude F ormat 6-29
Phase F ormat 6-30
Group Delay F ormat 6-30
Smith Chart F ormat 6-31
P olar F ormat 6-32
Linear Magnitude F ormat 6-33
SWR F ormat 6-33
Real F ormat 6-34
Imaginary F ormat 6-34
Group Delay Principles 6-35
Scale Reference Menu 6-38
Electrical Delay 6-38
Display Menu 6-39
Dual Channel Mode 6-40
Dual Channel Mode with Decoupled Channel P ower 6-40
Memory Math Functions 6-41
A djusting the Colors of the Display 6-41
Setting Display Intensity 6-41
Setting Default Colors 6-42
Blanking the Display 6-42
Saving Modi ed Colors 6-42
Recalling Modi ed Colors 6-42
The Modify Colors Menu 6-42
A veraging Menu 6-44
A veraging 6-44
Smoothing 6-45
IF Bandwidth Reduction 6-45
Markers 6-47
Marker Menu 6-48
Delta Mode Menu 6-48
Fixed Marker Menu 6-48
Marker Function Menu 6-49
Marker Search Menu 6-49
T arget Menu 6-49
Marker Mode Menu 6-49
P olar Marker Menu 6-49
Smith Marker Menu 6-49
Measurement Calibration 6-50
What is A ccuracy Enhancement? 6-50
What Causes Measurement Errors? 6-51
Directivity 6-51
Source Match 6-52
Load Match 6-52
Isolation (Crosstalk) 6-53
Frequency Response (Tracking) 6-53
Characterizing Microwave Systematic Errors 6-53
One-P ort Error Model 6-53
Device Measurement 6-59
Two-P ort Error Model 6-59
Contents-8

Calibration Considerations 6-65
Measurement P arameters 6-65
Device Measurements 6-65
Omitting Isolation Calibration 6-65
Saving Calibration Data 6-65
The Calibration Standards 6-66
Frequency Response of Calibration Standards 6-66
Electrical O set 6-67
Fringe Capacitance 6-67
How E ective Is A ccuracy Enhancement? 6-69
Correcting for Measurement Errors 6-71
Ensuring a V alid Calibration 6-71
Interpolated Error-correction 6-72
The Calibrate Menu 6-73
Response Calibration 6-73
Response and Isolation Calibration 6-73
One-P ort Calibration 6-73
Full Two-P ort Calibration 6-73
TRL*/LRM* Two-P ort Calibration 6-74
Restarting a Calibration 6-75
Cal Kit Menu 6-75
The Select Cal Kit Menu 6-75
Modifying Calibration Kits 6-76
De nitions 6-76
Procedure 6-76
Modify Calibration Kit Menu 6-77
De ne Standard Menus 6-78
Specify O set Menu 6-80
Label Standard Menu 6-81
Specify Class Menu 6-81
Label Class Menu 6-83
Label Kit Menu 6-83
V erify performance 6-84
TRL*/LRM* Calibration 6-85
Why Use TRL Calibration? 6-85
TRL T erminology 6-85
How TRL*/LRM* Calibration W orks 6-86
TRL* Error Model 6-86
Isolation 6-87
Source match and load match 6-88
Improving Raw Source Match and Load Match F or TRL*/LRM* Calibration . . 6-88
The TRL Calibration Procedure 6-90
Requirements for TRL Standards 6-90
F abricating and de ning calibration standards for TRL/LRM 6-91
TRL Options 6-93
P ower Meter Calibration 6-95
Primary Applications 6-95
Calibrated P ower Level 6-95
Compatible Sweep Types 6-95
Loss of P ower Meter Calibration Data 6-96
Interpolation in P ower Meter Calibration 6-96
P ower Meter Calibration Modes of Operation 6-96
Continuous Sample Mode (Each Sweep) 6-96
Sample-and-Sweep Mode (One Sweep) 6-97
P ower Loss Correction List 6-98
and S · Contents-9

P ower Sensor Calibration F actor List 6-98
Speed and A ccuracy 6-98
T est Equipment Used 6-98
Stimulus P arameters 6-98
Alternate and Chop Sweep Modes 6-100
Alternate 6-100
Chop 6-100
Calibrating for Non-Insertable Devices 6-101
A dapter Removal 6-101
Matched A dapters 6-101
Modify the Cal Kit Thru De nition 6-101
Using the Instrument State Functions 6-102
HP-IB Menu 6-103
Key 6-103
HP-IB ST A TUS Indicators 6-104
System Controller Mode 6-104
T alker/Listener Mode 6-104
P ass Control Mode 6-104
A ddress Menu 6-104
Using the P arallel P ort 6-105
The Copy Mode 6-105
The GPIO Mode 6-105
The System Menu 6-106
The Limits Menu 6-106
Edit Limits Menu 6-107
Edit Segment Menu 6-107
O set Limits Menu 6-108
Knowing the Instrument Modes 6-109
Network Analyzer Mode 6-109
External Source Mode 6-109
Primary Applications 6-109
Typical T est Setup 6-110
External Source Mode In-Depth Description 6-110
External Source A uto 6-110
External Source Manual 6-110
CW Frequency Range in External Source Mode 6-111
Compatible Sweep Types 6-111
External Source Requirements 6-111
Capture Range 6-111
Locking onto a signal with a frequency modulation component 6-111
Tuned Receiver Mode 6-111
Frequency O set Menu 6-112
Primary Applications 6-112
Typical T est Setup 6-112
Frequency O set In-Depth Description 6-113
The Receiver Frequency 6-113
The O set Frequency (LO) 6-113
Frequency Hierarchy 6-113
Frequency Ranges 6-113
Compatible Instrument Modes and Sweep Types 6-113
Receiver and Source Requirements 6-114
Display Annotations 6-114
Error Message 6-114
Spurious Signal P assband Frequencies 6-114
· Contents-10

Harmonic Operation (Option 002 only) 6-115
Typical T est Setup 6-115
Single-Channel Operation 6-115
Dual-Channel Operation 6-115
Coupling P ower Between Channels 1 and 2 6-115
Frequency Range 6-116
A ccuracy and input power 6-116
Time Domain Operation (Option 010) 6-117
The Transform Menu 6-117
General Theory 6-118
Time Domain Bandpass 6-119
A djusting the Relative V elocity F actor 6-119
Re ection Measurements Using Bandpass Mode 6-119
Interpreting the bandpass re ection response horizontal axis 6-120
Interpreting the bandpass re ection response vertical axis 6-120
Transmission Measurements Using Bandpass Mode 6-121
Interpreting the bandpass transmission response horizontal axis 6-121
Interpreting the bandpass transmission response vertical axis 6-121
Time domain low pass 6-122
Setting frequency range for time domain low pass 6-122
Minimum allowable stop frequencies 6-123
Re ection Measurements In Time Domain Low P ass 6-123
Interpreting the low pass response horizontal axis 6-123
Interpreting the low pass response vertical axis 6-123
F ault Location Measurements Using Low P ass 6-123
Transmission Measurements In Time Domain Low P ass 6-125
Measuring small signal transient response using low pass step 6-125
Interpreting the low pass step transmission response horizontal axis . . . 6-126
Interpreting the low pass step transmission response vertical axis 6-126
Measuring separate transmission paths through the test device using low
pass impulse mode 6-126
Time Domain Concepts 6-127
Masking 6-127
Windowing 6-128
Range 6-130
Resolution 6-131
Response resolution 6-131
Range resolution 6-132
Gating 6-133
Setting the gate 6-133
Selecting gate shape 6-134
Transforming CW Time Measurements Into the Frequency Domain 6-134
F orward Transform Measurements 6-135
Interpreting the forward transform vertical axis 6-135
Interpreting the forward transform horizontal axis 6-135
Demodulating the results of the forward transform 6-135
F orward transform range 6-136
T est Sequencing 6-138
In-Depth Sequencing Information 6-138
F eatures That Operate Di erently When Executed In a Sequence 6-138
Commands That Sequencing Completes Before the Next Sequence Command
Begins 6-138
Commands That Require a Clean Sweep 6-139
F orward Stepping In Edit Mode 6-139
Titles 6-139
Contents-11

Sequence Size 6-139
Embedding the V alue of the Loop Counter In a Title 6-139
A utostarting Sequences 6-139
The GPIO Mode 6-139
The Sequencing Menu 6-140
Gosub Sequence Command 6-140
TTL I/O Menu 6-140
TTL Output for Controlling P eripherals 6-140
TTL Input Decision Making 6-140
TTL Out Menu 6-142
Sequencing Special Functions Menu 6-142
Sequence Decision Making Menu 6-142
Decision Making Functions 6-142
Decision making functions jump to a softkey location, not to a speci c
sequence title 6-142
Having a sequence jump to itself 6-142
TTL input decision making 6-142
Limit test decision making 6-142
Loop counter decision making 6-143
Naming Files Generated by a Sequence 6-143
HP-GL Considerations 6-143
Entering HP-GL Commands 6-143
Special Commands 6-144
Entering Sequences Using HP-IB 6-144
Reading Sequences Using HP-IB 6-144
Ampli er T esting 6-145
Ampli er parameters 6-145
Gain Compression 6-146
Metering the power level 6-148
Mixer T esting 6-149
Frequency O set 6-149
Tuned Receiver 6-149
Mixer P arameters That Y ou Can Measure 6-150
A ccuracy Considerations 6-150
Attenuation at Mixer P orts 6-151
Filtering 6-152
Frequency Selection 6-153
LO Frequency A ccuracy and Stability 6-153
Up-Conversion and Down-Conversion De nition 6-153
Conversion Loss 6-156
Isolation 6-156
LO F eedthru / LO to RF Leakage 6-156
RF F eedthru 6-157
SWR / Return Loss 6-157
Conversion Compression 6-158
Phase Measurements 6-158
Amplitude and Phase Tracking 6-159
Phase Linearity and Group Delay 6-159
Connection Considerations 6-161
A dapters 6-161
Fixtures 6-162
If Y ou W ant to Design Y our Own Fixture 6-162
Reference Documents 6-163
General Measurement and Calibration T echniques 6-163
Fixtures and Non-Coaxial Measurements 6-163
Contents-12

On-W afer Measurements 6-164
Dynamic Range 7-1
HP 8753D Network Analyzer Speci cations 7-2
HP 8753D (50 ) with 7 mm T est P orts 7-2
Measurement P ort Characteristics 7-2
Transmission Measurement Uncertainties 7-3
Re ection Measurement Uncertainties 7-4
HP 8753D (50 ) with Type-N T est P orts 7-5
Measurement P ort Characteristics 7-5
Transmission Measurement Uncertainties 7-6
Re ection Measurement Uncertainties 7-7
HP 8753D (50 ) with 3.5 mm T est P orts 7-8
Measurement P ort Characteristics 7-8
Transmission Measurement Uncertainties 7-9
Re ection Measurement Uncertainties 7-10
HP 8753D (75 ) with Type-N T est P orts 7-11
Measurement P ort Characteristics 7-11
Transmission Measurement Uncertainties 7-12
Re ection Measurement Uncertainties 7-13
HP 8753D (75 ) with Type-F T est P orts 7-14
Measurement P ort Characteristics 7-14
Transmission Measurement Uncertainties 7-15
Re ection Measurement Uncertainties 7-16
Transmission Measurement Uncertainties 7-18
Re ection Measurement Uncertainties 7-19
Transmission Measurement Uncertainties 7-21
Re ection Measurement Uncertainties 7-22
Transmission Measurement Uncertainties 7-24
Re ection Measurement Uncertainties 7-25
Instrument Speci cations 7-26
HP 8753D Network Analyzer General Characteristics 7-33
Measurement Throughput Summary 7-33
Remote Programming 7-33
Interface 7-33
Transfer F ormats 7-33
Interface Function Codes 7-33
Front P anel Connectors 7-34
Probe P ower 7-34
Rear P anel Connectors 7-34
External Reference Frequency Input (EXT REF INPUT) 7-34
High-Stability Frequency Reference Output (10 MHz)(Option 001) 7-34
External A uxiliary Input (A UX INPUT) 7-34
External AM Input (EXT AM) 7-34
External Trigger (EXT TRIGGER) 7-35
T est Sequence Output (TEST SEQ) 7-35
Limit T est Output (LIMIT TEST) 7-35
T est P ort Bias Input (BIAS CONNECT) 7-35
Video Output (EXT MON) 7-35
HP-IB 7-35
P arallel P ort 7-36
RS-232 7-36
DIN Keyboard 7-36
Line P ower 7-36
Contents-13

Environmental Characteristics 7-36
General Conditions 7-36
Operating Conditions 7-36
Non-Operating Storage Conditions 7-36
W eight 7-37
Cabinet Dimensions 7-37
Internal Memory 7-37
Where to Look for More Information 9-1
Guide T erms and Conventions 9-2
Analyzer Functions 9-2
Cross Reference of Key Function to Programming Command 9-48
Softkey Locations 9-69
Where to Look for More Information 10-1
Error Messages in Alphabetical Order 10-2
Error Messages in Numerical Order 10-26
Where to Look for More Information 11-1
Measurement A ccessories A vailable 11-1
Calibration Kits 11-1
V eri cation Kit 11-2
HP 85029B 7 mm V eri cation Kit 11-2
T est P ort Return Cables 11-2
HP 11857D 7 mm T est P ort Return Cable Set 11-2
HP 11857B 75 Ohm Type-N T est P ort Return Cable Set 11-2
A dapter Kits 11-2
HP 11852B 50 to 75 Ohm Minimum Loss P ad 11-2
Transistor T est Fixtures 11-3
HP 11600B and 11602B Transistor Fixtures 11-3
HP 11608A Option 003 Transistor Fixture 11-3
HP 11858A Transistor Fixture A dapter 11-3
System A ccessories A vailable 11-4
System Cabinet 11-4
System T estmobile 11-4
Plotters and Printers 11-4
These plotters are compatible: 11-4
These printers are compatible: 11-4
Mass Storage 11-5
HP-IB Cables 11-5
Interface Cables 11-5
Keyboards 11-6
Controller 11-6
Sample Software 11-6
External Monitors 11-6
Connecting P eripherals 11-7
Connecting the P eripheral Device 11-7
Con guring the Analyzer for the P eripheral 11-8
If the P eripheral is a Printer 11-8
If the P eripheral is a Plotter 11-9
Contents-14

HPGL/2 Compatible Printer (used as a plotter) 11-9
P en Plotter 11-11
If the P eripheral is a P ower Meter 11-12
If the P eripheral is an External Disk Drive 11-12
If the P eripheral is a Computer Controller 11-12
Con guring the Analyzer to Produce a Time Stamp 11-13
HP-IB Programming Overview 11-14
HP-IB Operation 11-15
Device Types 11-15
T alker 11-15
Listener 11-15
Controller 11-15
HP-IB Bus Structure 11-16
Data Bus 11-16
Handshake Lines 11-16
Control Lines 11-16
HP-IB Requirements 11-17
HP-IB Operational Capabilities 11-18
HP-IB Status Indicators 11-19
Bus Device Modes 11-19
System-Controller Mode 11-20
T alker/Listener Mode 11-20
P ass-Control Mode 11-20
Setting HP-IB A ddresses 11-20
Analyzer Command Syntax 11-21
Code Naming Convention 11-21
V alid Characters 11-22
Units 11-23
HP-IB Debug Mode 11-23
Where to Look for More Information 12-1
Types of Memory and Data Storage 12-1
V olatile Memory 12-1
Non-V olatile Memory 12-2
Storing Data to Disk 12-4
Conserving Memory 12-5
Using Saved Calibration Sets 12-5
Preset State 12-6
A. The CITI le Data F ormat and K eyword Reference
The CITI le Data F ormat A -1
Description and Overview A -1
Data F ormats A -1
File and Operating System F ormats A -1
De nition of CITI le T erms A -2
A CITI le P ackage A -2
The CITI le Header A -2
An Array of Data A -2
CITI le Keyword A -3
CITI le Examples A -4
Example 2, An 8510 Display Memory File A -4
Example 3, 8510 Data le A -4
Example 4, 8510 3-T erm Frequency List Cal Set File A -5
Conclusion A -6
Contents-15

The CITI le Keyword Reference A -7
Index · Contents-16

1-1. HP 8753D Front P anel 1-4
1-2. Analyzer Display (Single Channel, Cartesian F ormat) 1-6
1-3. HP 8753D Rear P anel 1-10
2-1. Basic Measurement Setup 2-3
2-2. Example of Viewing Both Channels Simultaneously 2-5
2-3. Example Dual Channel With Split Display On 2-6
2-4. Example of a Display Title 2-8
2-5. A ctive Marker Control Example 2-10
2-6. A ctive and Inactive Markers Example 2-10
2-7. Marker 1 as the Reference Marker Example 2-11
2-8. Example of a Fixed Reference Marker Using
REF=FIXED MKR 2-12
2-9. Example of a Fixed Reference Marker Using
MKR ZERO 2-13
2-10. Example of Coupled and Uncoupled Markers 2-14
2-11. Example of a Log Marker in P olar F ormat 2-15
2-12. Example of Impedance Smith Chart Markers 2-16
2-13. Example of Setting the Start Frequency Using a Marker 2-17
2-14. Example of Setting the Stop Frequency Using a Marker 2-17
2-15. Example of Setting the Center Frequency Using a Marker 2-18
2-16. Example of Setting the Frequency Span Using Markerj 2-19
2-17. Example of Setting the Reference V alue Using a Marker 2-20
2-18. Example of Setting the Electrical Delay Using a Marker 2-21
2-19. Example of Searching for the Maximum Amplitude Using a Marker 2-22
2-20. Example of Searching for the Minimum Amplitude Using a Marker 2-23
2-21. Example of Searching for a T arget Amplitude Using a Marker 2-24
2-22. Example of Searching for a Bandwidth Using Markers 2-25
2-23. Example Statistics of Measurement Data 2-26
2-24. Device Connections for Measuring a Magnitude Response 2-27
2-25. Example Magnitude Response Measurement Results 2-28
2-26. Example Insertion Phase Response Measurement 2-28
2-27. Phase Samples 2-29
2-28. Device Connections for Measuring Electrical Length 2-30
2-29. Linearly Changing Phase 2-31
2-30. Example Best Flat Line with A dded Electrical Delay 2-32
2-31. Deviation From Linear Phase Example Measurement 2-33
2-32. Group Delay Example Measurement 2-34
2-33. Group Delay Example Measurement with Smoothing 2-34
2-34. Group Delay Example Measurement with Smoothing Aperture Increased . . . 2-35
2-35. Connections for SA W Filter Example Measurement 2-36
2-36. Example Flat Limit Line 2-38
2-37. Example Flat Limit Lines 2-39
2-38. Sloping Limit Lines 2-40
2-39. Example Single P oints Limit Line 2-41
2-40. Example Stimulus O set of Limit Lines 2-44
2-41. Diagram of Gain Compression 2-45
2-42. Gain Compression Using Linear Sweep and
D2/D1 to D2 ON 2-46
Figures · Contents-17

2-43. Gain Compression Using P ower Sweep 2-48
2-44. Gain and Reverse Isolation 2-50
2-45. Typical T est Setup for Tuned Receiver Mode 2-51
2-46. T est Sequencing Help Instructions 2-54
2-47. Fundamental and 2nd Harmonic P ower Levels in dBm 2-66
2-48. 2nd Harmonic P ower Level in dBc 2-67
2-49. Device Connections for Time Domain Transmission Example Measurement . . 2-68
2-50. Time Domain Transmission Example Measurement 2-69
2-51. Gating in a Time Domain Transmission Example Measurement 2-70
2-52. Gate Shape 2-71
2-53. Gating E ects in a Frequency Domain Example Measurement 2-72
2-54. Device Connections for Re ection Time Domain Example Measurement 2-73
2-55. Device Response in the Frequency Domain 2-74
2-56. Device Response in the Time Domain 2-75
3-1. Down Converter P ort Connections 3-3
3-2. Up Converter P ort Connections 3-3
3-3. R-Channel External Connection 3-4
3-4. An Example Spectrum of RF , LO , and IF Signals Present in a Conversion Loss
Measurement 3-7
3-5. Connections for R Channel and Source Calibration 3-8
3-6. Connections for a One-Sweep P ower Meter Calibration for Mixer Measurements 3-9
3-7. Diagram of Measurement Frequencies 3-10
3-8. Measurement Setup from Display 3-10
3-9. Conversion Loss Example Measurement 3-11
3-10. Connections for Broad Band P ower Meter Calibration 3-13
3-11. Connections for Receiver Calibration 3-14
3-12. Connections for a High Dynamic Range Swept IF Conversion Loss Measurement 3-15
3-13. Example of Swept IF Conversion Loss Measurement 3-16
3-14. Connections for a Response Calibration 3-18
3-15. Connections for a Conversion Loss Using the Tuned Receiver Mode 3-22
3-16. Example Fixed IF Mixer Measurement 3-23
3-17. Connections for a Group Delay Measurement 3-25
3-18. Group Delay Measurement Example 3-26
3-19. Conversion Loss and Output P ower as a Function of Input P ower Level Example 3-28
3-20. Connections for the First P ortion of Conversion Compression Measurement . . 3-29
3-21. Connections for the Second P ortion of Conversion Compression Measurement . 3-30
3-22. Measurement Setup Diagram Shown on Analyzer Display 3-31
3-23. Example Swept P ower Conversion Compression Measurement 3-32
3-24. Signal Flow in a Mixer Example 3-33
3-25. Connections for a Response Calibration 3-34
3-26. Connections for a Mixer Isolation Measurement 3-34
3-27. Example Mixer LO to RF Isolation Measurement 3-35
3-28. Connections for a Response Calibration 3-36
3-29. Connections for a Mixer RF F eedthrough Measurement 3-36
3-30. Example Mixer RF F eedthrough Measurement 3-37
4-1. Printer Connections to the Analyzer 4-3
4-2. Printing Two Measurements 4-7
4-3. P eripheral Connections to the Analyzer 4-8
4-4. Plot Components A vailable through De nition 4-12
4-5. Line Types A vailable 4-14
4-6. Locations of P1 and P2 in
SCALE PLOT [GRAT] Mode 4-15
4-7. Plot Quadrants 4-17
4-8. A utomatic File Naming Convention for LIF F ormat 4-19
4-9. Plot Filename Convention 4-26
4-10. Plotting Two Files on the Same P age 4-27
Contents-18

4-11. Plot Quadrants 4-28
4-12. Data Processing Flow Diagram 4-37
5-1. Standard Connections for a Response Error-Correction for Re ection
Measurement 5-9
5-2. Standard Connections for Response Error-Correction for Transmission
Measurements 5-10
5-3. Standard Connections for Receiver Calibration 5-11
5-4. Standard Connections for a Response and Isolation Error-Correction for
Re ection Measurements 5-14
5-5. Standard Connections for a Response and Isolation Error-Correction for
Transmission Measurements 5-16
5-6. Standard Connections for a One P ort Re ection Error-Correction 5-18
5-7. Standard Connections for Full Two port Error-Correction 5-20
5-8. Sample-and-Sweep Mode for P ower Meter Calibration 5-36
5-9. Continuous Correction Mode for P ower Meter Calibration 5-37
5-10. Calibrating for Noninsertable Devices 5-39
5-11. Noninsertable Device 5-41
5-12. A dapters Needed 5-42
5-13. Two-P ort Cal Set 1 5-43
5-14. Two-P ort Cal Set 2 5-44
5-15. Calibrated Measurement 5-45
6-1. Simpli ed Block Diagram of the Network Analyzer System 6-2
6-2. Data Processing Flow Diagram 6-4
6-3. A ctive Channel Keys 6-8
6-4. Entry Block 6-9
6-5. Stimulus Function Block 6-11
6-6. P ower Range Transitions in the A utomatic Mode 6-14
6-7. Response Function Block 6-25
6-8. S-P arameters of a Two-P ort Device 6-26
6-9. Re ection Impedance and A dmittance Conversions 6-28
6-10. Transmission Impedance and A dmittance Conversions 6-28
6-11. Log Magnitude F ormat 6-30
6-12. Phase F ormat 6-30
6-13. Group Delay F ormat 6-31
6-14. Standard and Inverse Smith Chart F ormats 6-32
6-15. P olar F ormat 6-32
6-16. Linear Magnitude F ormat 6-33
6-17. Typical SWR Display 6-33
6-18. Real F ormat 6-34
6-19. Constant Group Delay 6-35
6-20. Higher Order Phase Shift 6-35
6-21. Rate of Phase Change V ersus Frequency 6-36
6-22. V ariations in Frequency Aperture 6-36
6-23. Dual Channel Displays 6-40
6-24. E ect of A veraging on a Trace 6-44
6-25. E ect of Smoothing on a Trace 6-45
6-26. IF Bandwidth Reduction 6-46
6-27. Markers on Trace 6-47
6-28. Directivity 6-51
6-29. Source Match 6-52
6-30. Load Match 6-52
6-31. Sources of Error in a Re ection Measurement 6-54
6-32. Re ection Coecient 6-54
6-33. E ective Directivity E
6-55
6-34. Source Match E
6-55
Contents-19

6-35. Re ection Tracking E
6-56
6-36. \P erfect Load" T ermination 6-56
6-37. Measured E ective Directivity 6-57
6-38. Short Circuit T ermination 6-57
6-39. Open Circuit T ermination 6-58
6-40. Measured S
6-59
6-41. Major Sources of Error 6-60
6-42. Transmission Coecient 6-60
6-43. Load Match E
6-61
6-44. Isolation E
6-62
6-45. Full Two-P ort Error Model 6-63
6-46. Full Two-P ort Error Model Equations 6-64
6-47. Typical Responses of Calibration Standards after Calibration 6-68
6-48. Response versus S
1-P ort Calibration on Log Magnitude F ormat 6-69
6-49. Response versus S
1-P ort Calibration on Smith Chart 6-70
6-50. Response versus Full Two-P ort Calibration 6-70
6-51. HP 8753D functional block diagram for a 2-port error-corrected measurement
system 6-86
6-52. 8-term TRL error model and generalized coecients 6-87
6-53. Typical Measurement Set up 6-89
6-54. T est Setup for Continuous Sample Mode 6-97
6-55. T est Setup for Sample-and-Sweep Mode 6-97
6-56. Alternate and Chop Sweeps Overlaid 6-100
6-57. Instrument State Function Block 6-102
6-58. Typical Setup for the External Source Mode 6-110
6-59. Typical T est Setup for a Frequency O set Measurement 6-113
6-60. Typical Harmonic Mode T est Setup 6-115
6-61. Device Frequency Domain and Time Domain Re ection Responses 6-118
6-62. A Re ection Measurement of Two Cables 6-120
6-63. Transmission Measurement in Time Domain Bandpass Mode 6-121
6-64. Time Domain Low P ass Measurements of an Unterminated Cable 6-123
6-65. Simulated Low P ass Step and Impulse Response W aveforms (Real F ormat) . . . 6-124
6-66. Low P ass Step Measurements of Common Cable F aults (Real F ormat) 6-125
6-67. Time Domain Low P ass Measurement of an Ampli er Small Signal Transient
Response 6-126
6-68. Transmission Measurements Using Low P ass Impulse Mode 6-127
6-69. Masking Example 6-128
6-70. Impulse Width, Sidelobes , and Windowing 6-128
6-71. The E ects of Windowing on the Time Domain Responses of a Short Circuit . . 6-130
6-72. Response Resolution 6-132
6-73. Range Resolution of a Single Discontinuity 6-132
6-74. Sequence of Steps in Gating Operation 6-133
6-75. Gate Shape 6-134
6-76. Ampli er Gain Measurement 6-135
6-77. Combined E ects of Amplitude and Phase Modulation 6-136
6-78. Separating the Amplitude and Phase Components of T est-Device-Induced
Modulation 6-136
6-79. Range of a F orward Transform Measurement 6-137
6-80. P arallel P ort Input and Output Bus Pin Locations in GPIO Mode 6-141
6-81. Ampli er P arameters 6-145
6-82. Swept Frequency Ampli er Measurement of Absolute Fundamental, 2nd and 3rd
Harmonic Output Levels 6-145
6-83. Swept Frequency Ampli er Measurement of 2nd and 3rd Harmonic Distortion
(dBc) 6-146
6-84. Diagram of Gain Compression 6-147
Contents-20

6-85. Swept P ower Measurement of Ampli er's Fundamental Gain Compression and
2nd Harmonic Output Level 6-147
6-86. T est Con guration for Setting RF Input using A utomatic P ower Meter Calibration 6-148
6-87. Mixer P arameters 6-150
6-88. Conversion Loss versus Output Frequency Without Attenuators at Mixer P orts 6-151
6-89. Example of Conversion Loss versus Output Frequency Without Correct IF Signal
P ath Filtering 6-152
6-90. Example of Conversion Loss versus Output Frequency With Correct IF Signal
P ath Filtering and Attenuation at all Mixer P orts 6-152
6-91. Examples of Up Converters and Down Converters 6-153
6-92. Down Converter P ort Connections 6-154
6-93. Up Converter P ort Connections 6-155
6-94. Example Spectrum of RF , LO , and IF Signals Present in a Conversion Loss
Measurement 6-156
6-95. Main Isolation T erms 6-156
6-96. Conversion Loss and Output P ower as a Function of Input P ower Level 6-158
6-97. Connections for an Amplitude and Phase Tracking Measurement Between Two
Mixers 6-159
6-98. A dapter Considerations 6-161
7-1. External Trigger Circuit 7-35
11-1. P eripheral Connections to the Analyzer 11-7
11-2. HP-IB Bus Structure 11-16
11-3. Analyzer Single Bus Concept 11-19
Contents-21

0-1. Hewlett-P ackard Sales and Service Oces v
1-1. Comparing the HP 8753A/B/C/D 1-14
2-1. Connector Care Quick Reference 2-2
2-2. Gate Characteristics 2-71
4-1. Default V alues for Printing P arameters 4-6
4-2. Default P en Numbers and Corresponding Colors 4-13
4-3. Default P en Numbers for Plot Elements 4-13
4-4. Default Line Types for Plot Elements 4-14
4-5. Plotting P arameter Default V alues 4-16
4-6. HPGL Initialization Commands 4-23
4-7. HPGL T est File Commands 4-24
5-1. Di erences between PORT EXTENSIONS and ELECTRICAL DELA Y 5-3
5-2. Purpose and Use of Di erent Error-Correction Procedures 5-5
5-3. Typical Calibration Kit Standard and Corresponding Number 5-27
5-4. Characteristic P ower Meter Calibration Sweep Speed and A ccuracy 5-33
5-5. Band Switch P oints 5-50
5-6. Typical Recall State Times 5-58
6-1. Minimum Cycle Time (in seconds) 6-17
6-2. Display Colors with Maximum Viewing Angle 6-43
6-3. Calibration Standard Types and Expected Phase Shift 6-67
6-4. Standard De nitions 6-78
6-5. Standard Class Assignments 6-82
6-6. Characteristic P ower Meter Calibration Sweep Speed and A ccuracy 6-99
6-7. External Source Capture Ranges 6-111
6-8. Maximum Fundamental Frequency using Harmonic Mode 6-116
6-9. Time Domain Re ection F ormats 6-121
6-10. Minimum Frequency Ranges for Time Domain Low P ass 6-122
6-11. Impulse Width, Sidelobe Level, and Windowing V alues 6-129
6-12. Gate Characteristics 6-134
7-1. HP 8753D Dynamic Range 7-1
7-2. Measurement P ort Characteristics (Corrected
) for HP 8753D (50 ) with 7 mm
T est P orts 7-2
7-3. Measurement P ort Characteristics (Uncorrected
) for HP 8753D (50 ) with 7 mm
T est P orts 7-2
7-4. Measurement P ort Characteristics (Corrected)
for HP 8753D (50 ) with Type-N
T est P orts 7-5
7-5. Measurement P ort Characteristics (Corrected)
for HP 8753D (50 ) with 3.5 mm
T est P orts 7-8
7-6. Measurement P ort Characteristics (Corrected)
for HP 8753D (75 ) with Type-N
T est P orts 7-11
7-7. Measurement P ort Characteristics (Uncorrected)
for HP 8753D (75 ) with
Type-N T est P orts 7-11
7-8. Measurement P ort Characteristics (Corrected)
for HP 8753D (75 ) using HP
85039A F-M T est P orts 7-14
7-9. Measurement P ort Characteristics (Corrected)
for HP 8753D (75 ) using HP
85039A F-F T estports 7-17

7-10. Measurement P ort Characteristics (Corrected)
for HP 8753D (75 ) using HP
85039A M-M T estports 7-20
7-11. Measurement P ort Characteristics (Corrected)
for HP 8753D (75 ) using HP
85039A M-F T estports 7-23
9-1. Cross Reference of Key Function to Programming Command 9-48
9-2. Softkey Locations 9-70
11-1. Code Naming Convention 11-22
12-1. Memory Requirements of Calibration and Memory Trace Arrays 12-3
12-2. Sux Character De nitions 12-4
12-3. Preset Conditions (1 of 5) 12-7
12-4. P ower-on Conditions (versus Preset) 12-10
12-5. Results of P ower Loss to Non-V olatile Memory 12-11
* · * · Contents-23



HP 8753D Description and Options
This chapter contains information on the following topics:
Rear panel features and connectors
Service and support options
Changes between the HP 8753 network analyzers
Where to Look for More Information
A dditional information about many of the topics discussed in this chapter is located in the
following areas:
Chapter 2, \Making Measurements ," contains step-by-step procedures for making
measurements or using particular functions .
Chapter 4, \Printing, Plotting, and Saving Measurement Results ," contains instructions
for saving to disk or the analyzer internal memory , and printing and plotting displayed
measurements .
Chapter 5, \Optimizing Measurement Results ," describes techniques and functions for
achieving the best measurement results .
Chapter 6, \Application and Operation Concepts ," contains explanatory-style information
about many applications and analyzer operation.
HP 8753D Description and Options 1-1

The HP 8753D is a high performance vector network analyzer for laboratory or production
measurements of re ection and transmission parameters . It integrates a high resolution
synthesized RF source , an S-parameter test set, and a dual channel three-input receiver to
measure and display magnitude , phase , and group delay responses of active and passive RF
networks .
Two independent display channels and a large screen color display show the measured results
of one or both channels , in rectangular or polar/Smith chart formats .
F or information on options , refer to \Options A vailable" later in this chapter .
The analyzer has the additional following features:
Measurement functions selection with front panel keys and softkey menus .
External keyboard compatibility that allows you to title les and control the analyzer .
Internal automation, using test sequencing to program analyzer measurements and control
other devices without an external controller .
T est system automation with the addition of an external controller . This allows all of the
analyzer's measurement capabilities to be programmed over the Hewlett-P ackard Interface
Bus (HP-IB). (Refer to Chapter 11, \Compatible P eripherals"or the HP 8753D Network
Analyzer Programming Guide.)
A general purpose input/output (GPIO) bus that can control eight output bits and read ve
input bits through test sequencing. This can be useful for interfacing to material handlers
or custom test sets .
A utomatic sweep time that selects the minimum sweep time for the given IF bandwidth,
number of points , averaging mode , frequency range , and sweep type .
Built-in service diagnostics are available to simplify troubleshooting procedures .
Measurement exibility through trace math, data averaging, trace smoothing, electrical
delay , and accuracy enhancement.
External source mode capability that allows you to phase lock the analyzer's receiver to an
external source . Refer to Chapter 6, \Applications and Operation Concepts ."
Tuned receiver mode that allows you to use the receiver as a stand-alone device . Refer to
Chapter 6, \Applications and Operation Concepts ."
Complete re ection and transmission measurements in either 50 or 75 ohm impedance
environments .
Receiver/source frequency o set mode that allows you to set the analyzer's receiver and
source with a xed frequency o set for mixer test applications .
1-2 HP 8753D Description and Options
Analyzer Description · Control · P erformance

A ccuracy enhancement methods that range from normalizing data to complete one or two
port vector error correction with up to 1601 measurement points , and TRL*/LRM*. (V ector
error correction reduces the e ects of system directivity , frequency response , source and
load match, and crosstalk.)
P ower meter calibration that allows you to use an HP-IB compatible power meter to
monitor and correct the analyzer's output power at each data point. (The analyzer stores a
power correction table that contains the correction values .)
Printing, Plotting, and Saving
Direct print or plot output of displayed measurement results , with a time stamp if desired,
to a compatible peripheral with a serial, parallel, or HP-IB interface .
Instrument states storage in internal memory for the following times , or on disk
inde nitely .
T emperature at 70
C : 250 days (0.68 year) characteristically
T emperature at 40
C : 1244 days (3.4 years) characteristically
T emperature at 25
C : 10 years characteristically
LIF/DOS disk formats for saving instrument states and measurement data.
Integration of a high capacity micro- oppy disk drive .
HP 8753D Description and Options 1-3

Front P anel F eatures

Figure 1-1. HP 8753D Front P anel
Figure 1-1 shows the location of the following front panel features and key function blocks .
These features are described in more detail later in this chapter , and in Chapter 9, \Key
De nitions ."
labels . The display is divided into speci c information areas , illustrated in Figure 1-2.
source's frequency , power , and other stimulus functions .
and display functions of the active display channel.
keys allow you to select the active channel. Then any function you enter applies to this
active channel.
1-4 HP 8753D Description and Options

keys , and the number
pad. These allow you to enter numerical data and control the markers .
Y ou can use the numeric keypad to select digits , decimal points , and a minus sign for
numerical entries . Y ou must also select a units terminator to complete value inputs .
channel-independent system functions such as the following:
copying, save/recall, and HP-IB controller mode
frequency o set mode
harmonic measurements (Option 002)
time domain transform (Option 010)
HP-IB ST A TUS indicators are also included in this block.
key . This key returns the instrument to either a known factory preset state , or
a user preset state that can be de ned. Refer to Chapter 12, \Preset State and Memory
Allocation," for a complete listing of the instrument preset condition.
signals from a device under test. PORT 1 allows you to measure S
allows you to measure S
to an active probe for in-circuit measurements of ac circuits .
analyzer's R channel, for frequency o set mode .
measurement results for later analysis .
HP 8753D Description and Options 1-5


Figure 1-2. Analyzer Display (Single Channel, Cartesian F ormat)
The analyzer display shows various measurement information:
The grid where the analyzer plots the measurement data.
The currently selected measurement parameters .
The measurement data traces .
Figure 1-2 illustrates the locations of the di erent information labels described below .
In addition to the full-screen display shown in Figure 1-2, a split display is available , as
described in Chapter 2, \Making Measurements ." In the split display mode , the analyzer
provides information labels for each half of the display .
Several display formats are available for di erent measurements , as described under
" in Chapter 9, \Key De nitions ."
The start frequency of the source in frequency domain measurements .
The start time in CW mode (0 seconds) or time domain measurements .
The lower power value in power sweep .
When the stimulus is in center/span mode , the center stimulus value is shown in this
space .
1-6 HP 8753D Description and Options
Analyzer Display · \

The stop frequency of the source in frequency domain measurements .
The stop time in time domain measurements or CW sweeps .
The upper limit of a power sweep .
When the stimulus is in center/span mode , the span is shown in this space . The stimulus
values can be blanked, as described under \
FREQUENCY BLANK Key" in Chapter 9, \Key
De nitions ."
(F or CW time and power sweep measurements , the CW frequency is displayed centered
between the start and stop times or power values .)
channel.
The following notations are used:
A vg = Sweep-to-sweep averaging is on. The averaging count is shown immediately
below . (See \
Key" in Chapter 9, \Key De nitions .")
Cor = Error correction is on. (F or error-correction procedures , refer to Chapter 5,
\Optimizing Measurement Results ." F or error correction theory , refer to Chapter
6, \Application and Operation Concepts .")
C? = Stimulus parameters have changed from the error-corrected state , or interpolated
error correction is on. (F or error-correction procedures , refer to Chapter 5,
\Optimizing Measurement Results ." F or error correction theory , refer to Chapter
6, \Application and Operation Concepts .")
C2 = Full two-port error-correction is active and either the power range for each port
is di erent (uncoupled), or the
TESTSET SW HOLD is activated. The annotation
occurs because the analyzer does not switch between the test ports every sweep
under these conditions . The measurement stays on the active port after an
initial cycling between the ports . (The active port is determined by the selected
measurement parameter .) Y ou can update all the parameters by pressing
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MEASURE RESTART ,or
key .
Del = Electrical delay has been added or subtracted, or port extensions are active .
(See Chapter 6, \Application and Operation Concepts"and \
Chapter 9, \Key De nitions .")
ext = W aiting for an external trigger .
Ofs = Frequency o set mode is on. (See \Frequency O set Operation" in Chapter 6,
\Application and Operation Concepts .")
Of? = Frequency o set mode error , the IF frequency is not within 10 MHz of expected
frequency . LO inaccuracy is the most likely cause . (See \Frequency O set
Operation" in Chapter 6, \Application and Operation Concepts .")
Gat = Gating is on (time domain Option 010 only). (F or time domain measurement
procedures , refer to Chapter 2, \Making Measurements ." F or time domain theory ,
refer to Chapter 6 \Application and Operation Concepts .")
HP 8753D Description and Options 1-7
· Key" in

H=2 = Harmonic mode is on, and the second harmonic is being measured (harmonics
Option 002 only). See \Analyzer Options A vailable" later in this chapter .)
H=3 = Harmonic mode is on, and the third harmonic is being measured (harmonics
Option 002 only). (See \Analyzer Options A vailable" later in this chapter .)
Hold sweep . (See
HOLD in Chapter 9, \Key De nitions .")
man = W aiting for manual trigger .
PC = P ower meter calibration is on. (F or power meter calibration procedures , refer
to Chapter 5, \Optimizing Measurement Results ." F or power meter calibration
theory , refer to Chapter 6, \Application and Operation Concepts .")
PC? = The analyzer's source could not be set to the desired level, following a power
meter calibration. (F or power meter calibration procedures , refer to Chapter 5,
\Optimizing Measurement Results ." F or power meter calibration theory , refer to
Chapter 6, \Application and Operation Concepts .")
P? = Source power is unleveled at start or stop of sweep . (Refer to the
HP 8753D Network Analyzer Service Guide for troubleshooting.)
P# = Source power has been automatically set to minimum, due to receiver overload.
POWER in Chapter 9, \Key De nitions .")
PRm = P ower range is in manual mode .
Smo = Trace smoothing is on. (See \
" in Chapter 9, \Key De nitions ,")
tsH = Indicates that the test set hold mode is engaged.
That is , a mode of operation is selected which would cause repeated switching of
the step attenuator . This hold mode may be overridden. See
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NUMBER OF GROUPS in Chapter 9, \Key De nitions ."
" = F ast sweep indicator . This symbol is displayed in the status notation block when
sweep time is less than 1.0 second. When sweep time is greater than 1.0 second,
this symbol moves along the displayed trace .
sweep has been taken.
an attached keyboard or as described in Chapter 4, \Printing, Plotting, and Saving
Measurement Results ."
CHAN 1 and
CHAN 2 keys . If dual channel is on with an overlaid display , both channel 1
and channel 2 appear in this area.
measured, as selected using the
key . Also indicated in this area is the current
display memory status .
key .
key , in units
appropriate to the current measurement.
1-8 HP 8753D Description and Options

in polar formats , whichever you selected using the
key . The reference level is
also indicated by a small triangle adjacent to the graticule , at the left for channel 1 and at
the right for channel 2 in Cartesian formats .
current measurement. (Refer to \Using Analyzer Display Markers" in Chapter 2, \Making
Measurements .")
calculates when you access the menus with the
key . (Refer to \Using
Analyzer Display Markers" in Chapter 2, \Making Measurements .")
to the right of the analyzer display .
exceeded, and FAIL if any points exceed the limits .
HP 8753D Description and Options 1-9

Rear P anel F eatures and Connectors

Figure 1-3. HP 8753D Rear P anel
Figure 1-3 illustrates the features and connectors of the rear panel, described below .
Requirements for input signals to the rear panel connectors are provided in Chapter 7,
\Speci cations and Measurement Uncertainties ."
red, green, and blue video signals which you can use to drive an analog multi-sync
external monitor . The monitor must be compatible with the analyzer's 25.5 kHz scan rate
and video levels: 1 Vp-p , 0.7 V=white , 0 V=black, 0.3 V sync , sync on green.
compatible peripherals , and other instruments for an automated system. Refer to Chapter
11, \Compatible P eripherals ," in this document for HP-IB information, limitations , and
con gurations .
with a parallel input. Also included, is a general purpose input/output (GPIO) bus that
can control eight output bits and read ve input bits through test sequencing. Refer to
Chapter 11, \Compatible P eripherals ," for information on con guring a peripheral. Also
refer to \Application and Operation Concepts" for information on GPIO .
RS-232 (serial) input.
This provides a more convenient means to enter a title for storage les , as well as
substitute for the analyzer's front panel keyboard.
HP 8753D Network Analyzer Installation and Quick Start Guide or the
HP 8753D Network Analyzer Service Guide.
Analyzer Installation and Quick Start Guide.
1-10 HP 8753D Description and Options

input that can phase lock the analyzer to an external frequency standard for increased
frequency accuracy .
The analyzer automatically enables the external frequency reference feature when a
signal is connected to this input. When the signal is removed, the analyzer automatically
switches back to its internal frequency reference .
signal source , such as a detector or function generator , which you can then measure , using
the S-parameter menu. (Y ou can also use this connector as an analog output in service
routines , as described in the service manual.)
applied to the ALC circuitry of the analyzer's source . This input analog signal amplitude
modulates the RF output signal.
TTL-compatible signal that will trigger a measurement sweep . The trigger can be set to
external through softkey functions .
to be high or low , or pulse (10 seconds) high or low at the end of a sweep for robotic part
handler interface .
port 2. The fuses (1 A, 125 V) protect the port 1 and port 2 bias lines .
analyzer to an HP 85046A/B or 85047A S-parameter test set using the interconnect
cable supplied with the test set. The S-parameter test set is then fully controlled by the
analyzer .
HP 8753D Description and Options 1-11

Analyzer Options A vailable
Option 1D5, High Stability Frequency Reference
Option 1D5 o ers 0.05 ppm temperature stability from 0 to 60
C (referenced to 25
C).
Option 002, Harmonic Mode
Provides measurement of second or third harmonics of the test device's fundamental output
signal. Frequency and power sweep are supported in this mode . Harmonic frequencies can be
measured up to the maximum frequency of the receiver . However , the fundamental frequency
may not be lower than 16 MHz.
Option 006, 6 GHz Operation
Option 006 extends the maximum source and receiver frequency of the analyzer to 6 GHz.
Option 010, Time Domain
This option displays the time domain response of a network by computing the inverse F ourier
transform of the frequency domain response . It shows the response of a test device as a
function of time or distance . Displaying the re ection coecient of a network versus time
determines the magnitude and location of each discontinuity . Displaying the transmission
coecient of a network versus time determines the characteristics of individual transmission
paths . Time domain operation retains all accuracy inherent with the correction that is active in
the frequency domain. The time domain capability is useful for the design and characterization
of such devices as SA W lters , SA W delay lines , RF cables , and RF antennas .
Option 011, Receiver Con guration
Option 011 allows front panel access to the R, A, and B samplers and receivers . The transfer
switch, couplers , and bias tees have been removed. Therefore , external accessories are
required to make most measurements .
Option 075, 75 Impedance
Option 075 o ers 75 ohm impedance bridges with type-N test port connectors .
Option 1CM, Rack Mount Flange Kit Without Handles
Option 1CM is a rack mount kit containing a pair of anges and the necessary hardware to
mount the instrument, with handles detached, in an equipment rack with 482.6 mm (19 inches)
horizontal spacing.
Option 1CP , Rack Mount Flange Kit With Handles
Option 1CP is a rack mount kit containing a pair of anges and the necessary hardware to
mount the instrument with handles attached in an equipment rack with 482.6 mm (19 inches)
spacing.
1-12 HP 8753D Description and Options
·

Service and Support Options
The analyzer automatically includes a one-year on-site service warranty , where available . The
following service and support products are available with an HP 8753D network analyzer at
any time during or after the time of purchase . A dditional service and support options may be
available at some sites . Consult your local HP customer engineer for details .
On-Site System V eri cation (+23G)
On-site system veri cation (performed by a Hewlett-P ackard customer engineer), con rms
the system's error-corrected uncertainty performance by measuring traceable 7 mm devices .
It provides a hardcopy listing of both ideal and actual data, together with a certi cate of
traceability . Preventive maintenance is performed at the time of system veri cation. Travel
through Zone 3 (up to 100 miles/160 km from Hewlett-P ackard's nearest service-responsible
oce) is included.
Standard System Maintenance Service (+02A)
This option provides four-hour , on-site response through Travel Zone 3 on all service requests
for the HP 8753D (and a 50 ohm test set for Option 011), by a Hewlett-P ackard customer
engineer .
Basic System Maintenance Service (+02B)
This option provides next day on-site response through Travel Zone 3 on all service requests for
the HP 8753D (and a 50 ohm test set for Option 011), by a Hewlett-P ackard customer engineer .
Return to HP Full Service Agreement (+22A)
This option is a one-year service contract for any repair of the HP 8753D at a Hewlett-P ackard
repair facility . One complete calibration procedure is included.
Return to HP Repair Agreement (+22B)
This option provides repair of the HP 8753D at a Hewlett-P ackard repair facility for one year .
F ollowing repair , the instrument is tested functionally but is not fully calibrated.
Return to HP Calibration Agreement (+22C)
This option provides a once-a-year complete calibration procedure at a Hewlett-P ackard facility .
Return to HP Calibration (+22G)
This option is a one-time complete calibration procedure performed at a Hewlett-P ackard
facility . The procedure veri es that the HP 8753D is performing according to its published
speci cations .
HP 8753D Description and Options 1-13

Changes between the HP 8753 Network Analyzers
T able 1-1. Comparing the HP 8753A/B/C/D
F eature 8753A 8753B 8753C 8753D 8753D Opt
Fully integrated measurement system (built-in
No No No Y es No
T est port power range (dBm) y y y +10 to 85 y
A uto/manual power range selecting No No No Y es No
P ort power coupling/uncoupling No No No Y es No
Internal disk drive No No No Y es Y es
Precision frequency reference (Option 1D5) No No No Y es Y es
Frequency range - low end 300 kHz 300 kHz 300 kHz 30 kHz 30/300 kHz*
Ext. freq. range to 6 GHz (Option 006) No Y es Y es Y es Y es
75 system impedance (Option 075) y y y Y es y
TRL*/LRM* correction No No No Y es Y es
P ower meter calibration No Y es Y es Y es Y es
Interpolated error correction No Y es Y es Y es Y es
Max. error corrected measurement points 801 1601
Segmented error correction in freq. list mode No No Y es Y es Y es
Color CRT No No Y es Y es Y es
T est sequencing No Y es Y es Y es Y es
A utomatic sweep time No Y es Y es Y es Y es
External source capability No Y es Y es Y es Y es
Tuned receiver mode No Y es Y es Y es Y es
Printer/plotter bu er No Y es Y es Y es Y es
Harmonic measurements (Option 002) No Y es Y es Y es Y es
Frequency o set mode (mixer measurements) No Y es Y es Y es Y es
dc bias to test device y y y Y es y
No No No Y es Y es
User-de ned preset No No No Y es Y es
Non-volatile memory 16 Kbytes 16 Kbytes 16 Kbytes 512 Kbytes 512 Kbytes
30 kHz to 3 GHz 100 dB 100 dB 100 dB 110 dBz 100 dB
3 GHz to 6 GHz N/A 80 dB 80 dB 105 dB 110 dB
Real time clock No No No Y es Y es
* 300 kHz to 3 GHz, without Option 006, or 30 kHz to 6 GHz, with Option 006.
y F or this network analyzer , the feature is dependent on the test set being used.
z 90 dB from 30 kHz to 50 kHz, 100 dB from 300 kHz to 16 MHz.
1-14 HP 8753D Description and Options
test set) · keyboard · Dynamic range

This Chapter contains the following example procedures for making measurements or using
particular functions:
Basic measurement sequence and example
Magnitude and insertion phase response
Electrical length and phase distortion
Deviation from linear phase
Gain and reverse isolation
High P ower Measurements
Where to Look for More Information
A dditional information about many of the topics discussed in this Chapter is located in the
following areas:
Chapter 4, \Printing, Plotting, and Saving Measurement Results ," contains instructions
for saving to disk or the analyzer internal memory , and printing and plotting displayed
measurements .
Chapter 5, \Optimizing Measurement Results ," describes techniques and functions for
achieving the best measurement results .
Chapter 6, \Application and Operation Concepts ," contains explanatory-style information
about many applications and analyzer operation.
Chapter 9, \Key De nitions ," describes all the front panel keys and softkeys .
Chapter 11, \Compatible P eripherals ," lists measurement and system accessories , and other
applicable equipment compatible with the analyzer .

Principles of Microwave Connector Care
Proper connector care and connection techniques are critical for accurate , repeatable
measurements .
Refer to the calibration kit documentation for connector care information. Prior to making
connections to the network analyzer , carefully review the information about inspecting,
cleaning and gaging connectors .
Having good connector care and connection techniques extends the life of these devices . In
addition, you obtain the most accurate measurements .
This type of information is typically located in Chapter 3 of the calibration kit manuals .
F or additional connector care instruction, contact your local Hewlett-P ackard Sales and Service
Oce about course numbers HP 85050A+24A and HP 85050A+24D .
See the following table for quick reference tips about connector care .
T able 2-1. Connector Care Quick Reference
Keep connectors clean T ouch mating-plane surfaces
Extend sleeve or connector nut Set connectors contact-end down
Use plastic end-caps during storage
Inspect all connectors carefully Use a damaged connector - ever
Look for metal particles , scratches , and dents
Try compressed air rst Use any abrasives
Use isopropyl alcohol Get liquid into plastic support beads
Clean and zero the gage before use Use an out-of-spec connector
Use the correct gage type
Use correct end of calibration block
Gage all connectors before rst use
Align connectors carefully Apply bending force to connection
Make preliminary connection lightly Over tighten preliminary connection
Turn only the connector nut Twist or screw any connection
Use a torque wrench for nal connect Tighten past torque wrench \break" point

Basic Measurement Sequence and Example
There are ve basic steps when you are making a measurement.
Damage may result to the device under test if it is sensitive to analyzer's
default output power level. T o avoid damaging a sensitive DUT , perform step 2
before step 1.
This example procedure shows you how to measure the transmission response of a bandpass
lter .

Figure 2-1. Basic Measurement Setup
.
T o set preset to \F actory Preset," press:
PRESET: FACTORY
PRESET
Setting the Frequency Range .
M/

M/
Y ou could also press the
STOP
keys and enter the frequency range
limits as start frequency and stop frequency values .
Setting the Source P ower .
POWER
MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Y ou could also press
POWER RANGES and select one of the
power ranges to keep the power setting within the de ned range .
Setting the Measurement.
NUMBER OF POINTS
MENU
Trans:FWD S21 (B/R)
NNNNNNNNNNNNNNNNNNNNNNNNNNNN
SCALE REF
AUTOSCALE
measurement errors .
SELECT DISK
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
INTERNAL MEMORY
NNNNNNNNNNNNNNNNNNNN
RETURN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SAVE STATE
SA VE/RECALL
M/
NNNNNNNNNNNNNN
COPY
PLOT )
Refer to Chapter 4, \Printing, Plotting, and Saving Measurement Results ," for procedures
on how to de ne a print, plot, or save . F or information on con guring a peripheral, refer
to Chapter 11, \Compatible P eripherals ."

Using the Display Functions
T o View Both Measurement Channels
In some cases , you may want to view more than one measured parameter at a time .
Simultaneous gain and phase measurements for example , are useful in evaluating stability in
negative feedback ampli ers . Y ou can easily make such measurements using the dual channel
display .
DISPLA Y
DUAL CHAN ON

Figure 2-2. Example of Viewing Both Channels Simultaneously
· · Making Measurements 2-5

Y ou can view the measurements on separate displays , press:
MORE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SPLIT DISP ON The
analyzer shows channel 1 on the upper half of the display and channel 2 on the lower half
of the display . The analyzer also defaults to measuring S
on channel 1 and S

Figure 2-3. Example Dual Channel With Split Display On
T o return to one display , press:
SPLIT DISPLAY OFF
Y ou can control the stimulus functions of the two channels independent of each
other , by pressing
COUPLED CH OFF .
T o Save a Data Trace to the Display Memory
DISPLA Y
DATA ! MEMORY to store the current active measurement data in the memory of
the active channel.
The data trace is now also the memory trace . Y ou can use a memory trace for subsequent math
manipulations .
T o View the Measurement Data and Memory Trace
The analyzer default setting shows you the current measurement data for the active channel.
MEMORY
DISPLA Y
This is the only memory display mode where you can change the smoothing and gating of
the memory trace .
DISPLA Y

T o Divide Measurement Data by the Memory Trace
Y ou can use this feature for ratio comparison of two traces , for example , measurements of gain
or attenuation.
Save a Data Trace to the Display Memory ."
NNNNNNNNNNNNNNNNNNNNNNNNN
DISPLA Y
DATA/MEM to divide the data by the memory .
The analyzer normalizes the data to the memory , and shows the results .
T o Subtract the Memory Trace from the Measurement Data Trace
Y ou can use this feature for storing a measured vector error , for example , directivity . Then,
you can later subtract it from the device measurement.
Save a Data Trace to the Display Memory ."
DATA-MEM to subtract the memory from the measurement data.
The analyzer performs a vector subtraction on the complex data.
T o Ratio Measurements in Channel 1 and 2
Y ou may want to use this feature when making ampli er measurements to produce a trace that
represents gain compression. F or example , with the channels uncoupled, you can increase the
power for channel 2 while channel 1 remains unchanged. This will allow you to observe the
gain compression on channel 2 .
MENU
COUPLED CH OFF to uncouple the channels .
a.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MENU
NUMBER OF POINTS and notice the number of points setting, shown
on the analyzer display .
b.
MENU
NUMBER OF POINTS and enter the same value that you observed
for the channel 1 setting.
DUAL CHAN ON
NNNNNNNNNNNNNN
MORE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
D2/D1 TO D2 ON to ratio channels 1 and 2, and put
the results in the channel 2 data array . This ratio is applied to the complex data.
compression point.

T o Title the A ctive Channel Display
NNNNNNNNNNNNNNNNN
TITLE to access the title menu.
MORE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ERASE TITLE and enter the title you want for your measurement display .
If you have a DIN keyboard attached to the analyzer , type the title you want from the
keyboard. Then press
to enter the title into the analyzer . Y ou can enter a title
that has a maximum of 50 characters .
If you do not have a DIN keyboard attached to the analyzer , enter the title from the
analyzer front panel.
a. Turn the front panel knob to move the arrow pointer to the rst character of the title .
b.
SELECT LETTER .
c. Repeat the previous two steps to enter the rest of the characters in your title . Y ou can
enter a title that has a maximum of 50 characters .
d. Press
DONE to complete the title entry .

Figure 2-4. Example of a Display Title

Using Analyzer Display Markers
The analyzer markers provide numerical readout of trace data. Y ou can control the marker
search, the statistical functions , and the capability for quickly changing stimulus parameters
with markers , from the
key .
Markers have a stimulus value (the x-axis value in a Cartesian format) and a response value
(the y-axis value in a Cartesian format). In a polar or Smith chart format, the second part of
a complex data pair is also provided as an auxiliary response value . When you switch on a
marker , and no other function is active , the analyzer shows the marker stimulus value in the
active entry area. Y ou can control the marker with the front panel knob , the step keys , or the
front panel numeric keypad.
If you activate both data and memory traces , the marker values apply to the data trace .
If you activate only the memory trace , the marker values apply to the memory trace .
If you activate a memory math function (data/memory or data-memory), the marker values
apply to the trace resulting from the memory math function.
The examples in this section are shown with lter measurement results . The measurement
parameters are set as follows:
CENTER
M/
SP AN
M/
T o Use Continuous and Discrete Markers
The analyzer can either place markers on discrete measured points , or move the markers
continuously along a trace by interpolating the data value between measured points .
MARKER MODE MENU and select one of the following choices:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
on the trace , by interpolating between measured points . This default mode allows you to
conveniently obtain round numbers for the stimulus value .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
trace points determined by the stimulus settings . This may be the best mode to use with
automated testing, using a computer or test sequencing because the analyzer does not
interpolate between measured points .
functions when the value entered in a search or positioning function does not
exist as a measurement point.

T o A ctivate Display Markers
T o switch on marker 1 and make it the active marker , press:
MARKER 1
NNNNNNNNNNNNNNNNNNNNNNNNN
MARKER
The active marker appears on the analyzer display as r . The active marker stimulus value is
displayed in the active entry area. Y ou can modify the stimulus value of the active marker ,
using the front panel knob or numerical keypad. All of the marker response and stimulus
values are displayed in the upper right corner of the display .

Figure 2-5. A ctive Marker Control Example
T o switch on the corresponding marker and make it the active marker , press:
NNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER 2 ,
MARKER 3 ,
MARKER 4 , or
All of the markers , other than the active marker , become inactive and are represented on the
analyzer display as .

Figure 2-6. A ctive and Inactive Markers Example
· · 2-10 Making Measurements

T o switch o all of the markers , press:
ALL OFF
T o Use Delta () Markers
This is a relative mode , where the marker values show the position of the active marker
relative to the delta reference marker . Y ou can switch on the delta mode by de ning one of the
ve markers as the delta reference .
NNNNNNNNNNNNNNNNNNNNNNN
MODE MENU
REF=1 to make marker 1 a reference marker .
a. Turn the front panel knob .
a. Enter the frequency value (relative to the reference marker) on the numeric keypad.
MARKER 2 and move marker 2 to any position that you want to measure in reference
to marker 1.

Figure 2-7. Marker 1 as the Reference Marker Example
NNNNNNNNNNNNNNNNNNNNNNN
MODE MENU
REF=2
T o A ctivate a Fixed Marker
When a reference marker is xed, it does not rely on a current trace to maintain its xed
position. The analyzer allows you to activate a xed marker with one of the following key
sequences:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER
MODE MENU
REF=FIXED MKR
NNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER
MKR ZERO

REF=FIXED MKR K ey to activate a Fixed Reference Marker
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER
MODE MENU
REF=FIXED MKR
MODE MENU
FIXED MKR POSITION
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FIXED MKR STIMULUS and turn the front panel knob or enter a value from the front panel
keypad.
The marker is shown on the display as a small delta (), smaller than the inactive marker
triangles .
FIXED MKR VALUE and turn the front panel knob or enter a value from the front panel
keypad.
In a Cartesian format the setting is the y-axis value . In polar or Smith chart format, with a
magnitude/phase marker , a real/imaginary marker , an R+jX marker , or a G+jB marker , the
setting applies to the rst part of the complex data pair . (Fixed marker response values are
always uncoupled in the two channels .)
format, press:
FIXED MKR AUX VALUE and turn the front panel knob or enter a value from the front panel
keypad.
This value is the second part of complex data pair , and applies to a magnitude/phase marker ,
a real/imaginary marker , an R+jX marker , or a G+jB marker . (Fixed marker auxiliary
response values are always uncoupled in the two channels .)

Figure 2-8.
Example of a Fixed Reference Marker Using
REF=FIXED MKR

MKR ZERO K ey to A ctivate a Fixed Reference Marker
Marker zero enters the position of the active marker as the reference position. Alternatively ,
you can specify the xed point with
FIXED MKR POSITION . Marker zero is canceled by
switching delta mode o .
and turn the front panel knob or enter a value from the front panel keypad.
you set in the previous step , press:
MKR ZERO and turn the front panel knob or enter a value from the front panel keypad.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MODE MENU
FIXED MKR POSITION
FIXED MKR STIMULUS and turn the front panel knob
or enter a value from the front panel keypad.

Figure 2-9. Example of a Fixed Reference Marker Using
MKR ZERO
Using the · · Making Measurements 2-13

T o Couple and Uncouple Display Markers
At a preset state , the markers have the same stimulus values on each channel, but they can be
uncoupled so that each channel has independent markers .
MARKER F CTN
MARKER MODE MENU and select from the following keys:
for the two display channels .
values for the two display channels . This allows you to control the marker stimulus values
independently for each channel.

Figure 2-10. Example of Coupled and Uncoupled Markers
T o Use P olar F ormat Markers
The analyzer can display the marker value as magnitude and phase , or as a real/imaginary pair:
NNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNN
LIN MKR gives linear magnitude and phase ,
LOG MKR gives log magnitude and phase ,
gives the real value rst, then the imaginary value .
Y ou can use these markers only when you are viewing a polar display format. (The format is
available from the
key .)
F or greater accuracy when using markers in the polar format, it is
recommended to activate the discrete marker mode . Press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MKR MODE MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKERS:DISCRETE .
F ORMA T
POLAR
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER F CTN
MARKER MODE MENU
POLAR MKR MENU

LIN MKR if you want to view the magnitude and the phase of the active marker .
The magnitude values appear in units and the phase values appear in degrees .
LOG MKR if you want to view the logarithmic magnitude and the phase of the
active marker . The magnitude values appear in dB and the phase values appear in
degrees .
Re/Im MKR if you want to view the real and imaginary pair , where the complex
data is separated into its real part and imaginary part. The analyzer shows the rst
marker value the real part (M cos ), and the second value is the imaginary part
(M sin , where M=magnitude).

Figure 2-11. Example of a Log Marker in P olar F ormat
T o Use Smith Chart Markers
The amount of power re ected from a device is directly related to the impedance of the device
and the measuring system. Each value of the re ection coecient () uniquely de nes a device
impedance; = 0 only occurs when the device and analyzer impedance are exactly the same .
The re ection coecient for a short circuit is: = 1
. Every other value for also
corresponds uniquely to a complex device impedance , according to the equation:
= [( 1 + ) / (1 )]Z
is your test device impedance and Z
is the measuring system's characteristic
impedance .
F or greater accuracy when using markers in the Smith chart format, it is
recommended to activate the discrete marker mode . Press
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MKR MODE MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKERS:DISCRETE .
F ORMA T
SMITH CHART .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER F CTN
MARKER MODE MENU
SMITH MKR MENU and turn the front panel knob or
enter a value from the front panel keypad to read the resistive and reactive components of
the complex impedance at any point along the trace . This is the default Smith chart marker .

The marker annotation tells that the complex impedance is capacitive in the bottom half of
the Smith chart display and is inductive in the top half of the display .
LIN MKR if you want the analyzer to show the linear magnitude and the phase of
the re ection coecient at the marker .
LOG MKR if you want the analyzer to show the logarithmic magnitude and the
phase of the re ection coecient at the active marker . This is useful as a fast method of
obtaining a reading of the log magnitude value without changing to log magnitude format.
Re/Im MKR if you want the analyzer to show the values of the re ection
coecient at the marker as a real and imaginary pair .
R+
X MKR to show the real and imaginary parts of the device impedance at
the marker . Also shown is the equivalent series inductance or capacitance (the series
resistance and reactance , in ohms).
G+
B MKR to show the complex admittance values of the active marker in
rectangular form. The active marker values are displayed in terms of conductance (in
Siemens), susceptance , and equivalent parallel circuit capacitance or inductance . Siemens
are the international unit of admittance and are equivalent to mhos (the inverse of ohms).

Figure 2-12. Example of Impedance Smith Chart Markers
T o Set Measurement P arameters Using Markers
The analyzer allows you to set measurement parameters with the markers , without going
through the usual key sequence . Y ou can change certain stimulus and response parameters to
make them equal to the current active marker value .

Setting the Start Frequency
and turn the front panel knob or enter a value from the front panel
keypad to position the marker at the value that you want for the start frequency .
MARKER !START to change the start frequency value to the value of the active marker .

Figure 2-13. Example of Setting the Start Frequency Using a Marker
Setting the Stop Frequency
and turn the front panel knob or enter a value from the front panel
keypad to position the marker at the value that you want for the stop frequency .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER !STOP to change the stop frequency value to the value of the active marker .

Figure 2-14. Example of Setting the Stop Frequency Using a Marker
Press · Press · Making Measurements 2-17

Setting the Center Frequency
and turn the front panel knob or enter a value from the front panel
keypad to position the marker at the value that you want for the center frequency .
MARKER ! CENTER to change the center frequency value to the value of the active
marker .

Figure 2-15. Example of Setting the Center Frequency Using a Marker
Press · 2-18 Making Measurements

Setting the Frequency Span
Y ou can set the span equal to the spacing between two markers . If you set the center
frequency before you set the frequency span, you will have a better view of the area of
interest.
NNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNN
MODE MENU
REF=1
MARKER 2 .
markers where you want the frequency span.
Iterate between marker 1 and marker 2 by pressing
NNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER 2 , respectively ,
and turning the front panel knob or entering values from the front panel keypad to position
the markers around the center frequency . When nished positioning the markers , make sure
that marker 2 is selected as the active marker .
NNNNNNNNNNNNNNNNNNNNNNNNNN
Step 2 can also be performed using
MARKER 1 . However , when
using this method, it will not be possible to iterate between marker zero and
marker 1.
MARKER ! SPAN to change the frequency span to the range between
marker 1 and marker 2.

Figure 2-16. Example of Setting the Frequency Span Using Markerj

Setting the Display Reference V alue
and turn the front panel knob or enter a value from the front panel
keypad to position the marker at the value that you want for the analyzer display reference
value .
MARKER ! REFERENCE to change the reference value to the value of the active marker .

Figure 2-17. Example of Setting the Reference V alue Using a Marker
2-20 Making Measurements

Setting the Electrical Delay
This feature adds phase delay to a variation in phase versus frequency , therefore it is only
applicable for ratioed inputs .
F ORMA T
PHASE .
and turn the front panel knob or enter a value from the front panel
keypad to position the marker at a point of interest.
MARKER !DELAY to automatically add or subtract enough line length to the receiver
input to compensate for the phase slope at the active marker position. This e ectively
attens the phase trace around the active marker . Y ou can use this to measure the electrical
length or deviation from linear phase .
A dditional electrical delay adjustments are required on devices without constant group delay
over the measured frequency span.

Figure 2-18. Example of Setting the Electrical Delay Using a Marker
Setting the CW Frequency
and either turn the front panel knob or enter the value , followed by
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNN
SEQ
SPECIAL FUNCTIONS
MKR ! CW .
Y ou can use this function to set the marker to a gain peak in an ampli er . After pressing
MKR ! CW FREQ , activate a CW frequency power sweep to look at the gain compression with
increasing input power .

T o Search for a Speci c Amplitude
These functions place the marker at an amplitude-related point on the trace . If you switch on
tracking, the analyzer searches every new trace for the target point.
Searching for the Maximum Amplitude
MARKER F CTN
MKR SEARCH to access the marker search menu.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
trace .

Figure 2-19.
Example of Searching for the Maximum Amplitude Using a Marker

Searching for the Minimum Amplitude
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER F CTN
MKR SEARCH to access the marker search menu.
trace .

Figure 2-20.
Example of Searching for the Minimum Amplitude Using a Marker

Searching for a T arget Amplitude
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER F CTN
MKR SEARCH to access the marker search menu.
trace .
enter the new value from the front panel keypad.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEARCH RIGHT .

Figure 2-21.
Example of Searching for a T arget Amplitude Using a Marker

Searching for a Bandwidth
The analyzer can automatically calculate and display the 3 dB bandwidth (BW:), center
frequency (CENT:), Q, and loss of the device under test at the center frequency . (Q stands for
\quality factor ," de ned as the ratio of a circuit's resonant frequency to its bandwidth.) These
values are shown in the marker data readout.
and turn the front panel knob or enter a value from the front panel keypad
to place the marker at the center of the lter passband.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MKR ZERO
MARKER F CTN
MKR SEARCH to access the marker search menu.
WIDTHS ON to calculate the center stimulus value , bandwidth, and the Q of a
bandpass or band reject shape on the measurement trace .
WIDTH VALUE and enter the new value from the front panel keypad.

Figure 2-22. Example of Searching for a Bandwidth Using Markers
Tracking the Amplitude that Y ou are Searching
Speci c Amplitude ."
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER F CTN
MKR SEARCH
TRACKING ON to track the speci ed amplitude search
with every new trace and put the active marker on that point.
When tracking is not activated, the analyzer nds the speci ed amplitude on the current
sweep and the marker remains at same stimulus value , regardless of changes in the trace
response value with subsequent sweeps .

T o Calculate the Statistics of the Measurement Data
This function calculates the mean, standard deviation, and peak-to-peak values of the section
of the displayed trace between the active marker and the delta reference . If there is no delta
reference , the analyzer calculates the statistics for the entire trace .
NNNNNNNNNNNNNNNNNNNNNNN
MARKER
MODE MENU
REF=1 to make marker 1 a reference marker .
Turn the front panel knob .
Enter the frequency value on the numeric keypad.
MARKER 2 and move marker 2 to any position that you want to measure in reference
to marker 1.
NNNNNNNNNNNNNNNNNNNNNNNNNN
MKR MODE MENU
STATS ON to calculate and view the mean, standard
deviation, and peak-to-peak values of the section of the measurement data between the
active marker and the delta reference marker .
An application for this feature is to nd the peak-to-peak value of passband ripple without
searching separately for the maximum and minimum values .
If you are viewing a measurement in the polar or Smith Chart format, the analyzer
calculates the statistics using the rst value of the complex pair (magnitude , real part,
resistance , or conductance).

Figure 2-23. Example Statistics of Measurement Data

Measuring Magnitude and Insertion Phase Response
The analyzer allows you to make two di erent measurements simultaneously . Y ou can make
these measurements in di erent formats for the same parameter . F or example , you could
measure both the magnitude and phase of transmission. Y ou could also measure two di erent
).
This measurement example shows you how to measure the maximum amplitude of a SA W lter
and then how to view the measurement data in the phase format, which provides information
about the phase response .
Measuring the Magnitude Response

Figure 2-24. Device Connections for Measuring a Magnitude Response
and choose the measurement settings . F or this example the measurement
parameters are set as follows:
MEAS
M/
SP AN
M/
NNNNNNNNNNNNNNNNN
MENU
POWER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SCALE REF
AUTO SCALE
CHAN 2
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MEAS
SCALE REF
AUTO SCALE
Y ou may also want to select settings for the number of data points , averaging, and IF
bandwidth.
channel 2.
NNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNN
CAL
CALIBRATE MENU
RESPONSE
THRU .
N
NNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNN
THRU .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AUTO SCALE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER F CTN
MKR SEARCH
SEARCH: MAX

Figure 2-25. Example Magnitude Response Measurement Results
Measuring Insertion Phase Response
press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DISPLA Y
DUAL CHAN ON
N
NNNNNNNNNNNNNNNN
PHASE
F ORMA T
The channel 2 portion of Figure 2-26 shows the insertion phase response of the device under
test. The analyzer measures and displays phase over the range of 180
. As phase
changes beyond these values , a sharp 360
transition occurs in the displayed data.

Figure 2-26. Example Insertion Phase Response Measurement

The phase response shown in Figure 2-27 is undersampled; that is , there is more than
phase delay between frequency points . If the = >180
, incorrect phase and delay
information may result. Figure 2-27 shows an example of phase samples being with less
and greater than 180

Figure 2-27. Phase Samples
Undersampling may arise when measuring devices with long electrical length. T o correct this
problem, the frequency span should be reduced, or the number of points increased until is
per point. Electrical delay may also be used to compensate for this e ect (as
shown in the next example procedure).

Measuring Electrical Length and Phase Distortion
The analyzer mathematically implements a function similar to the mechanical \line stretchers"
of earlier analyzers . This feature simulates a variable length lossless transmission line , which
you can add to or remove from the analyzer's receiver input to compensate for interconnecting
cables , etc . In this example , the electronic line stretcher measures the electrical length of a
SA W lter .
The analyzer allows you to measure the linearity of the phase shift through a device over a
range of frequencies and the analyzer can express it in two di erent ways:
deviation from linear phase
Figure 2-28. Device Connections for Measuring Electrical Length
and choose the measurement settings . F or this example , the measurement
settings include reducing the frequency span to eliminate under sampled phase response .
Press the following keys as shown:
MEAS
M/
SP AN
M/
NNNNNNNNNNNNNNNNN
MENU
POWER
NNNNNNNNNNNNNNNNN
F ORMA T
PHASE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SCALE REF
AUTO SCALE
Y ou may also want to select settings for the number of data points , averaging, and IF
bandwidth.

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNN
CAL
CALIBRATE MENU
RESPONSE
THRU
AUTO SCALE
SCALE REF
a 2 MHz span. Other lters may require a wider frequency span to see the e ects of phase
shift.
The linearly changing phase is due to the device's electrical length. Y ou can measure this
changing phase by adding electrical length (electrical delay) to compensate for it.

Figure 2-29. Linearly Changing Phase
and turn the front panel knob or enter a value from the front panel keypad.
MARKER F CTN
MARKER ! DELAY
This function calculates and adds in the appropriate electrical delay by taking a 10%
span about the marker , measuring the , and computing the delay as the negative of
/frequency .

ELECTRICAL DELAY and turn the front panel knob to increase the
SCALE REF
electrical length until you achieve the best at line , as shown in Figure 2-30.
The measurement value that the analyzer displays represents the electrical length of your
device relative to the speed of light in free space . The physical length of your device is
related to this value by the propagation velocity of its medium.
V elocity factor is the ratio of the velocity of wave propagation in a coaxial
cable to the velocity of wave propagation in free space . Most cables have a
relative velocity of about 0.66 the speed in free space . This velocity depends
on the relative permittivity of the cable dielectric (
V el ocity F actor =
is the relative permittivity of the cable dielectric
Y ou could change the velocity factor to compensate for propagation velocity by
MORE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
VELOCITY FACTOR (enter the value)
. This will help
the analyzer to accurately calculate the equivalent distance that corresponds to
the entered electrical delay .

Figure 2-30. Example Best Flat Line with A dded Electrical Delay
SCALE REF
ELECTRICAL DELAY
In this example , there is a large amount of electrical delay due to the long electrical length
of the SA W lter under test.
This portion of the example shows you how to measure the linearity of the phase shift over
a range of frequencies . The analyzer allows you to measure this linearity and read it in two
di erent ways: deviation from linear phase , or group delay .
Deviation From Linear Phase
By adding electrical length to \ atten out" the phase response , you have removed the linear
phase shift through your device . The deviation from linear phase shift through your device is
all that remains .

SCALE REF
SCALE DIV and turn the front panel knob or enter a value from the front panel
keypad.
phase , press:
NNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER F CTN
MKR MODE MENU
STATS ON
It is possible to use delta markers to measure peak-to-peak deviation in only
one portion of the trace , see \T o Calculate the Statistics of the Measurement
Data" located earlier in this Chapter .

Figure 2-31. Deviation From Linear Phase Example Measurement
The phase linearity of many devices is speci ed in terms of group or envelope delay .
The analyzer can translate this information into a related parameter , group delay . Group
delay is the transmission time through your device under test as a function of frequency .
Mathematically , it is the derivative of the phase response which can be approximated by the
following ratio:
where is the di erence in phase at two frequencies separated by F . The quantity F is
commonly called the \aperture" of the measurement. The analyzer calculates group delay from
its phase response measurements .
The default aperture is the total frequency span divided by the number of points across the
display (i.e . 201 points or 0.5% of the total span in this example).
\Deviation from Linear Phase ."
NNNNNNNNNNNNNNNNN
DELAY
SCALE REF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SCALE DIV

and turn the front panel knob or enter a value from the front panel keypad.

Figure 2-32. Group Delay Example Measurement
Group delay measurements may require a speci c aperture (f) or frequency spacing between
measurement points . The phase shift between two adjacent frequency points must be less than
, otherwise incorrect group delay information may result.
approximately 1% of the frequency span, press:
SMOOTHING ON .
When you increase the aperture , the analyzer removes ne grain variations from the
response . It is critical that you specify the group delay aperture when you compare group
delay measurements .

Figure 2-33. Group Delay Example Measurement with Smoothing

points over which the analyzer calculates the group delay , press:
SMOOTHING APERTURE
As the aperture is increased the \smoothness" of the trace improves markedly , but at the
expense of measurement detail.

Figure 2-34.
Group Delay Example Measurement with Smoothing Aperture Increased
5 · · Making Measurements 2-35

T esting A Device with Limit Lines
Limit testing is a measurement technique that compares measurement data to constraints that
you de ne . Depending on the results of this comparison, the analyzer will indicate if your
device either passes or fails the test.
Limit testing is implemented by creating individual at, sloping, and single point limit lines on
the analyzer display . When combined, these lines can represent the performance parameters
for your device under test. The limit lines created on each measurement channel are
independent of each other .
This example measurement shows you how to test a bandpass lter using the following
procedures:
creating at limit lines
creating sloping limit lines
creating single point limit lines
running a limit test
Setting Up the Measurement P arameters
Figure 2-35. Connections for S A W Filter Example Measurement
and choose the measurement settings . F or this example the measurement
settings are as follows:
MEAS
M/
SP AN
M/
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SCALE REF
AUTO SCALE
Y ou may also want to select settings for the number of data points , power , averaging, and IF
bandwidth.
NNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNN
CAL
CALIBRATE MENU
RESPONSE
THRU

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AUTO SCALE
Creating Flat Limit Lines
In this example procedure , the following at limit line values are set:
Frequency Range : P ower Range
127 MHz to 140 MHz : 27 dB to 21 dB
100 MHz to 123 MHz : 200 dB to 65 dB
146 MHz to 160 MHz : 200 dB to 65 dB
The minimum value for measured data is 200 dB .
LIMIT MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LIMIT LINE ON
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
EDIT LIMIT LINE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CLEAR LIST
NNNNNNNNNNN
YES
SYSTEM
ADD
The analyzer generates a new segment that appears on the center of the display .
STIMULUS VALUE
M/
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
UPPER LIMIT
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LOWER LIMIT
NNNNNNNNNNNNNN
DONE
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Y ou could also set the upper and lower limits by using the
DELTA LIMITS keys . T o use these keys for the entry , press:
MIDDLE VALUE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DELTA LIMITS
This would correspond to a test speci cation of 24 3 dB .
LIMIT TYPE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FLAT LINE
NNNNNNNNNNNNNNNNNNNN
RETURN

ADD
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
STIMULUS VALUE
M/
NNNNNNNNNNNNNN
DONE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LIMIT TYPE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SINGLE POINT
NNNNNNNNNNNNNNNNNNNN
RETURN
Figure 2-36 shows the at limit lines that you have just created with the following
parameters:
stimulus from 127 MHz to 140 MHz
upper limit of 21 dB
lower limit of 27 dB

Figure 2-36. Example Flat Limit Line
ADD
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
STIMULUS VALUE
M/
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
UPPER LIMIT
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LOWER LIMIT
NNNNNNNNNNNNNN
DONE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNN
LIMIT TYPE
FLAT LINE
RETURN
NNNNNNNNNNN
ADD
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
STIMULUS VALUE
M/
NNNNNNNNNNNNNN
DONE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LIMIT TYPE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SINGLE POINT
NNNNNNNNNNNNNNNNNNNN
RETURN

ADD
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
STIMULUS VALUE
M/
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
UPPER LIMIT
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LOWER LIMIT
NNNNNNNNNNNNNN
DONE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNN
LIMIT TYPE
FLAT LINE
RETURN
NNNNNNNNNNN
ADD
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
STIMULUS VALUE
M/
NNNNNNNNNNNNNN
DONE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNN
LIMIT TYPE
SINGLE POINT
RETURN

Figure 2-37. Example Flat Limit Lines
Creating a Sloping Limit Line
This example procedure shows you how to make limits that test the shape factor of a SA W
lter . The following limits are set:
Frequency Range : P ower Range
123 MHz to 125 MHz : 65 dB to 26 dB
144 MHz to 146 MHz : 26 dB to 65 dB
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNN
SYSTEM
LIMIT MENU
LIMIT LINE ON
EDIT LIMIT LINE
CLEAR LIST
YES
the lter , press:
ADD
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
STIMULUS VALUE
M/
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
UPPER LIMIT
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LOWER LIMIT
NNNNNNNNNNNNNN
DONE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LIMIT TYPE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SLOPING LINE
NNNNNNNNNNNNNNNNNNNN
RETURN

ADD
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
STIMULUS VALUE
M/
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
UPPER LIMIT
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LOWER LIMIT
NNNNNNNNNNNNNN
DONE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNN
LIMIT TYPE
SINGLE POINT
RETURN
the lter , press:
ADD
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
STIMULUS VALUE
M/
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
UPPER LIMIT
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LOWER LIMIT
NNNNNNNNNNNNNN
DONE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNN
LIMIT TYPE
SLOPING LINE
RETURN
ADD
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
STIMULUS VALUE
M/
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
UPPER LIMIT
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LOWER LIMIT
NNNNNNNNNNNNNN
DONE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LIMIT TYPE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SINGLE POINT
NNNNNNNNNNNNNNNNNNNN
RETURN
Y ou could use this type of limit to test the shape factor of a lter .

Figure 2-38. Sloping Limit Lines

Creating Single P oint Limits
In this example procedure , the following limits are set:
from 23 dB to 28.5 dB at 141 MHz
from 23 dB to 28.5 dB at 126.5 MHz
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNN
SYSTEM
LIMIT MENU
LIMIT LINE ON
EDIT LIMIT LINE
CLEAR LIST
YES
downward pointing, indicating the upper test limit
upward pointing, indicating the lower test limit
Press:
ADD
NNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
STIMULUS VALUE
M/
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
UPPER LIMIT
N
LOWER LIMIT
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
28.5
DONE
NNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LIMIT TYPE
SINGLE POINT
NNNNNNNNNNNNNNNNNNNN
RETURN
NNNNNNNNNNN
ADD
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
STIMULUS VALUE
126.5
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
UPPER LIMIT
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LOWER LIMIT
28.5
DONE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LIMIT TYPE
SINGLE POINT
NNNNNNNNNNNNNNNNNNNN
RETURN

Figure 2-39. Example Single P oints Limit Line

This example shows you how to edit the upper limit of a limit line .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SYSTEM
LIMIT MENU
LIMIT LINE ON
EDIT LIMIT LINE
press:
SEGMENT
NNNNNNNNNNNNNNNNNNNNNN
* or
NNNNNNNNNNNNNNNNNNNNNNN
SEGMENT and enter the segment number followed by
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNN
EDIT
UPPER LIMIT
DONE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SYSTEM
LIMIT MENU
LIMIT LINE ON
EDIT LIMIT LINE
press:
SEGMENT
* or
NNNNNNNNNNNNNNNNNNNNNNN
SEGMENT and enter the segment number followed by
DELETE

Running a Limit T est
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SYSTEM
LIMIT MENU
LIMIT LINE ON
EDIT LIMIT LINE
Reviewing the Limit Line Segments
The limit table data that you have previously entered is shown on the analyzer display .
SEGMENT
* and
earlier in this section.
A ctivating the Limit T est
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SYSTEM
LIMIT MENU
LIMIT TEST ON
BEEP FAIL ON
Selecting the beep fail indicator
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
BEEP FAIL ON is optional and will add
approximately 50 ms of sweep cycle time . Because the limit test will still work
if the limits lines are o , selecting
LIMIT LINE ON is also optional.
The limit test results appear on the right side on the analyzer display . The analyzer indicates
whether the lter passes or fails the de ned limit test:
The message FAIL will appear on the right side of the display if the limit test fails .
The analyzer beeps if the limit test fails and if
BEEP FAIL ON has been selected.
The analyzer alternates a red trace where the measurement trace is out of limits .
A TTL signal on the rear panel BNC connector \LIMIT TEST" provides a pass/fail (5 V/0 V)
indication of the limit test results .

O setting Limit Lines
The limit o set functions allow you to adjust the limit lines to the frequency and output level
of your device . F or example , you could apply the stimulus o set feature for testing tunable
lters . Or , you could apply the amplitude o set feature for testing variable attenuators , or
passband ripple in lters with variable loss .
This example shows you the o set feature and the limit test failure indications that can appear
on the analyzer display .
press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LIMIT MENU
LIMIT LINE OFFSETS
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
STIMULUS OFFSET
M/
The analyzer beeps and a FAIL notation appears on the analyzer display , as shown in
Figure 2-40.

Figure 2-40. Example Stimulus O set of Limit Lines
STIMULUS OFFSET
AMPLITUDE OFFSET
The analyzer beeps and a FAIL notation appears on the analyzer display .
AMPLITUDE OFFSET

Gain compression occurs when the input power of an ampli er is increased to a level that
reduces the gain of the ampli er and causes a nonlinear increase in output power . The point at
which the gain is reduced by 1 dB is called the 1 dB compression point. The gain compression
will vary with frequency , so it is necessary to nd the worst case point of gain compression in
the frequency band.
Once that point is identi ed, you can perform a power sweep of that CW frequency to measure
the input power at which the 1 dB compression occurs and the absolute power out (in dBm) at
compression. The following steps provide detailed instruction on how to apply various features
of the analyzer to accomplish these measurements .

Figure 2-41. Diagram of Gain Compression
e ect of noise on the trace , press:
NNNNNNNNNNNNNNNNN
IF BW
NNNNNNNNNNNNNN
Measurement Results ," for instructions on how to make a measurement correction.
or step 6. (Step 5 uses trace math and step 6 uses uncoupled channels and the display
D1/D2 to D2 ON .)
NNNNNNNNNNNNNNNNNNNNNNNNNN
DISPLA Y
DATA ! MEMORY
DATA/MEM to produce a normalized trace .
a.
DISPLA Y
DUAL CHANNEL ON to view both channels simultaneously .

b. T o uncouple the channel stimulus so that the channel power will be uncoupled, press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
COUPLED CH OFF
This will allow you to separately increase the power for channel 2 and channel 1 ,
so that you can observe the gain compression on channel 2 while channel 1 remains
unchanged.
c. T o display the ratio of channel 2 data to channel 1 data on the channel 2 display , press:
NNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DISPLAY
MORE
D2/D1 to D2 ON
This produces a trace that represents gain compression only .
MARKER
MARKER 1 and position the marker at approximately mid-span.
SCALE REF
SCALE/DIV
to change the scale to 1 dB per division.
MENU
POWER .
using the step keys or the front panel knob .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER F CTN
MKR SEARCH
SEARCH:MIN

Figure 2-42. Gain Compression Using Linear Sweep and
D2/D1 to D2 ON
COUPLED CH OFF was selected, recouple the channel stimulus by pressing:
MENU
COUPLED CH ON
MARKER MODE MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKERS:DISCRETE
MARKER F CTN

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SPECIAL FUNCTIONS
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER ! CW
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MENU
SWEEP TYPE MENU
POWER SWEEP .
Now channel 1 is displaying a gain compression curve . (Do not pay attention to channel 2
at this time .)
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL
INTERPOL ON
CORRECTION ON
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CHAN 2
DISPLA Y
DUAL CHANNEL ON .
NNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
D2/D1 to D2 ON was selected, press
D2/D1 to D2 OFF .
NNNNN
MEAS
INPUT PORTS
B .
Now channel 2 displays absolute output power (in dBm) as a function of power input.
SCALE REF
SCALE/DIV
to change the scale of channel 2 to 10 dB per division.
to change the scale of channel 1 to 1 dB per division.
A receiver calibration will improve the accuracy of this measurement. Refer to
Chapter 5, \Optimizing Measurement Results ."
MARKER MODE MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKERS:COUPLED .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER F CTN
MKR SEARCH
SEARCH:MAX
NNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER
MKR ZERO
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEARCH:TARGET
MARKER F CTN
MKR SEARCH
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER
MKR MODE MENU
MARKERS:UNCOUPLED .
of the ampli er (in dBm), press:
MODE MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MODE OFF
MARKER


Figure 2-43. Gain Compression Using P ower Sweep
2-48 Making Measurements

Measuring Gain and Reverse Isolation Simultaneously
Since an ampli er will have high gain in the forward direction and high isolation in the reverse
direction, the gain (S
) will be much greater than the reverse isolation (S
). Therefore , the
power you apply to the input of the ampli er for the forward measurement (S
considerably lower than the power you apply to the output for the reverse measurement
). By applying low power in the forward direction, you'll prevent the ampli er from being
saturated. A higher power in the reverse direction keeps noise from being a factor in the
measurement and accounts for any losses caused by attenuators or couplers on the ampli er's
output needed to lower the output power into the analyzer . The following steps demonstrate
the features that best accomplish these measurements .
COUPLED CH ON .
Coupling the channels allows you to have the same frequency range and calibration applied
to channel 1 and channel 2 .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PORT POWER [UNCOUPLED] .
Uncoupling the port power allows you to apply di erent power levels at each port. In
Figure 2-44 , the port 1 power is set to 25 dBm for the gain measurement (S21) and the
port 2 power is set to 0 dBm for the reverse isolation measurement (S12).
NNNNNNNNNNNNNNNNN
CHAN 1
MEAS
POWER and set the power level for
port 1.
N
N
NNNNNNNNNNNNNNNN
POWER and set the power level for
MENU
port 2.
\Optimizing Measurement Results" Chapter for error-correction procedures .
DISPLA Y
DUAL CHAN ON .
Y ou can view both measurements simultaneously by using the dual channel display mode .
Refer to Figure 2-44. If the port power levels are in di erent power ranges , one of the
displayed measurements will not be continually updated and the annotation tsH will appear
on the left side of the display . Refer to \Source Attenuator Switch Protection"section in
Chapter 6, \Application and Operation Concepts ," for information on how to override this
state .

T o obtain best accuracy , you should set the power levels prior to performing
the calibration. However , the analyzer compensates for nominal power changes
you make during a measurement, so that the error correction still remains
approximately valid. In these cases , the Cor annunciator will change to C? .

Figure 2-44. Gain and Reverse Isolation
2-50 Making Measurements

Measurements Using the Tuned Receiver Mode
In the tuned receiver mode , the analyzer's receiver operates independently of any signal
source . This mode is not phase-locked and functions in all sweep types . The analyzer tunes
the receiver to a synthesized CW input signal at a precisely speci ed frequency . All phase
lock routines are bypassed, increasing sweep speed signi cantly . The external source must
be synthesized, and must drive the analyzer's external frequency reference . The analyzer's
internal source frequency is not accurate , and the internal source should not be used in the
tuned receiver mode .
Using the analyzer's tuned receiver mode is useful for automated test applications where an
external synthesized source is available and applications where speed is important. Although
the tuned receiver mode can function in all sweep types , it is typically used in CW applications .
A ctivate the tuned receiver mode by pressing
INSTRUMENT MODE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TUNED RECEIVER .
tuned receiver mode .

Figure 2-45. Typical T est Setup for Tuned Receiver Mode
Tuned receiver mode in-depth description
30 kHz to 3 GHz (6 GHz for Option 006)
All sweep types may be used.

An analyzer in tuned receiver mode can receive input signals into PORT 1, PORT 2, or R
CHANNEL IN.
Input power range speci cations are provided in Chapter 7, \Speci cations and Measurement
Uncertainties ."
External Source Requirements · 2-52 Making Measurements

T est sequencing allows you to automate repetitive tasks . As you make a measurement, the
analyzer memorizes the keystrokes . Later you can repeat the entire sequence by pressing a
single key . Because the sequence is de ned with normal measurement keystrokes , you do not
need additional programming expertise . Subroutines and limited decision-making increases the
exibility of test sequences . In addition, the GPIO outputs can be controlled in a test sequence ,
and the GPIO inputs can be tested in a sequence for conditional branching. F or in-depth
sequencing information, refer to \T est Sequencing" in Chapter 6, \Application and Operation
Concepts ."
The test sequence function allows you to create , title , save , and execute up to six independent
sequences internally .
Y ou can also save sequences to disk and transfer them between the analyzer and an external
computer controller .
The following procedures are based on an actual measurement example , that shows you how to
do the following:
change a sequence title
name les generated by a sequence
There are also three example sequences:

SEQ
NEW SEQ/MODIFY SEQ
As shown in Figure 2-46, a list of instructions appear on the analyzer display to help you
create or edit a sequence .

Figure 2-46. T est Sequencing Help Instructions
SEQUENCE 1 SEQ1
This choice selects sequence position #1. The default title is SEQ1 for this sequence .
Refer to \Changing the Sequence Title ," (located later in this Chapter) for information on
how to modify a sequence title .

F or this example , a SA W lter measurement is set up with the following parameters:
SELECT DISK
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
INTERNAL MEMORY
SA VE/RECALL
NNNNNNNNNNNNNNNNNNNN
RETURN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RECALL STATE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNN
LOG MAG
F ORMA T
CENTER
M/
SP AN
M/
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNN
SCALE REF
AUTOSCALE
The above keystrokes will create a displayed list as shown:
CENTER
SPAN
SCALE/DIV
AUTO SCALE
DONE SEQ MODIFY
SEQ
When you create a sequence , the analyzer stores it in volatile memory where
it will be lost if you switch o the instrument power (except for sequence #6
which is stored in the analyzer non-volatile memory). However , you may store
sequences to a oppy disk.
T o run a stored test sequence , press:
and the softkey labeled with desired sequence number
or , press:
SEQ
DO SEQUENCE and the softkey labeled with the desired sequence number .
T o stop a sequence before it has nished, press
.

SEQ
NEW SEQ/MODIFY SEQ
press:
SEQUENCE 1 SEQ1
* or
If you use the
* key to move the cursor through the list of commands , the commands are
actually performed when the cursor points to them. This feature allows the sequence to
be tested one command at a time .
If you wish to scroll through the sequence without executing each line as you do so , you
can press the
key and scroll through the command list backwards .
SEQ
DONE SEQ MODIFY to exit the modify (edit) mode .
SEQ
NEW SEQ/MODIFY SEQ
press:
SEQUENCE 1 SEQ1
want to insert a new command, by pressing:
* or
If you use the
* key to move the cursor through the list of commands , the commands are
actually performed when the cursor points to them. This feature allows the sequence to
be tested one command at a time .
If you wish to scroll through the sequence without executing each line as you do so , you
can press the
key and scroll through the command list backwards .
if you want to activate the averaging function, press:
A VG
AVERAGING ON
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQ
DONE SEQ MODIFY to exit the modify (edit) mode .

PRESET
SEQ
NEW SEQ/MODIFY SEQ
press:
SEQUENCE 1 SEQ1
The following list is the commands entered in \Creating a Sequence ." Notice that for longer
sequences , only a portion of the list can appear on the screen at one time .
CENTER
SPAN
SCALE/DIV
AUTO SCALE
cursor (! ) next to the command that you wish to modify , press:
* or
If you use the
* key to move the cursor through the list of commands , the commands are
actually performed when the cursor points to them. This feature allows the sequence to
be tested one command at a time .
If you wish to scroll through the sequence without executing each line as you do so , you
can press the
key and scroll through the command list backwards .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQ
DONE SEQ MODIFY to exit the modify (edit) mode .
Clearing a Sequence from Memory
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQ
MORE
CLEAR SEQUENCE

Changing the Sequence Title
If you are storing sequences on a disk, you should replace the default titles (SEQ1, SEQ2 . . . ).
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQ
MORE
TITLE SEQUENCE and select the particular sequence softkey .
The analyzer shows the available title characters . The current title is displayed in the
upper-left corner of the screen.
If you have an attached DIN keyboard, you can press
and then type the new lename .
If you do not have an attached DIN keyboard, press
ERASE TITLE and turn the front
panel knob to point to the characters of the new lename , pressing
you stop at each character .
The analyzer cannot accept a title ( le name) that is longer than eight characters . Y our titles
must also begin with a letter , and contain only letters and numbers .
NNNNNNNNNNNNN
T o complete the titling, press
DONE .
Naming Files Generated by a Sequence
The analyzer can automatically increment the name of a le that is generated by a sequence
using a loop structure . (See example ''Generating Files in a Loop Counter Example Sequence''
later in this chapter .)
T o access the sequence lename menu, press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FILE UTILITIES
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQUENCE FILENAMING
This menu presents two choices:
FILE NAME FILE0 supplies a name for the saved state and/or data le . This also brings up
the Title File Menu.
PLOT NAME PLOTFILE supplies a name for the plot le generated by a plot-to-disk command.
This also brings up the Title File Menu.
The above keys show the current lename in the 2nd line of the softkey .
When titling a le for use in a loop function, you are restricted to only 2 characters in the
lename due to the 6 character length of the loop counter keyword \[LOOP]." When the le is
actually written, the [LOOP] keyword is expanded to only 5 ASCII characters (digits), resulting
in a 7 character lename .
After entering the 2 character lename , press:
LOOP COUNTER
NNNNNNNNNNNNNN
DONE

Storing a Sequence on a Disk
MORE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
STORE SEQ TO DISK and select the particular sequence softkey .
The disk drive access light should turn on brie y . When it goes out, the sequence has been
saved.
The analyzer will overwrite a le on the disk that has the same title .
· · Making Measurements 2-59

Loading a Sequence from Disk
F or this procedure to work, the desired le must exist on the disk in the analyzer drive .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQ
MORE
LOAD SEQ FROM DISK
READ SEQ FILE TITLS
If the desired sequence is not among the rst six les , press:
READ SEQ FILE TITLS until the desired le name appears .
illuminate brie y .
If you know the title of the desired sequence , you can title the sequence (1-6)
with the name , and load the sequence . This is also how you can control the
sequence number of an imported titled sequence .
Purging a Sequence from Disk
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQ
MORE
STORE SEQ TO DISK
PURGE SEQUENCES
READ SEQ FILE TITLS
If the desired sequence is not among the rst six les , press:
READ SEQ FILE TITLS until the desired le name appears .
illuminate brie y .
P eripherals .")
NNNNNNNNNNNNNN
MORE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PRINT SEQUENCE and the softkey for the desired sequence .
If the sequence is on a disk, load the sequence (as described in a previous
procedure) and then follow the printing sequence .

Cascading Multiple Example Sequences
By cascading test sequences , you can create subprograms for a larger test sequence . Y ou can
also cascade sequences to extend the length of test sequences to greater than 200 lines .
In this example , you are shown two sequences that have been cascaded. Y ou can do this by
having the last command in sequence 1 call sequence position 2, regardless of the sequence
title . Because sequences are identi ed by position, not title , the call operation will always go to
the sequence loaded into the given position.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQ
NEW SEQ/MODIFY SEQ
SEQUENCE 1 SEQ1
CENTER
M/
SP AN
M/
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DO SEQUENCE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQUENCE 2
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DONE SEQ MODIFY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NEW SEQ/MODIFY SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQUENCE 2 SEQ2
N
MEAS
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
F ORMA T
LOG MAG
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SCALE REF
AUTOSCALE
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQ
DONE SEQ MODIFY
The following sequences will be created:
CENTER
SPAN
DO SEQUENCE
SEQUENCE 2
SEQUENCE SEQ2
SCALE/DIV
AUTO SCALE
Y ou can extend this process of calling the next sequence from the last line of the present
sequence to 6 internal sequences , or an unlimited number of externally stored sequences .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQUENCE 1 SEQ1

Loop Counter Example Sequence
This example shows you the basic steps necessary for constructing a looping structure within
a test sequence . A typical application of this loop counter structure is for repeating a speci c
measurement as you step through a series of CW frequencies or dc bias levels . F or an example
application, see \Fixed IF Mixer Measurements" in Chapter 3.
that you want to repeat, press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQ
NEW SEQ/MODIFY SEQ
SEQUENCE 1 SEQ1
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SPECIAL FUNCTIONS
DECISION MAKING
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LOOP COUNTER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQ
DO SEQUENCE
SEQUENCE 2
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DONE SEQ MODIFY
This will create a displayed list as shown:
LOOP COUNTER
DO SEQUENCE
SEQUENCE 2
T o create a second sequence that will perform a desired measurement function,
decrement the loop counter , and call itself until the loop counter value is equal to zero ,
press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQ
NEW SEQ/MODIFY SEQ
SEQUENCE 2 SEQ2
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MEAS
SCALE REF
AUTO SCALE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER F CTN
MKR SEARCH
SEARCH: MAX
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQ
SPECIAL FUNCTIONS
DECISION MAKING
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DECR LOOP COUNTER
IF LOOP COUNTER<> 0
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQUENCE 2 SEQ2
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQ
DONE SEQ MODIFY
This will create a displayed list as shown:
AUTO SCALE
SEARCH MAX
DECR LOOP COUNTER
IF LOOP COUNTER <> 0 THEN DO
SEQUENCE 2
T o run the loop sequence , press:
PRESET
SEQUENCE 1 SEQ1

Generating Files in a Loop Counter Example Sequence
This example shows how to increment the names of tiles that are generated by a sequence with
a loop structure .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQ
NEW SEQ/MODIFY SEQ
SEQUENCE 1 SEQ 1
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SPECIAL FUNCTIONS
DECISION MAKING
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LOOP COUNTER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SA VE/RECALL
SELECT DISK
INTERNAL DISK
NNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RETURN
DEFINE DISK-SAVE
DATA ONLY ON
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNN
LOCAL
SET ADDRESSES
PLOTTER PORT
DISK
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQ
DO SEQUENCE
SEQUENCE 2
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DONE SEQ MODIFY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NEW SEQ/MODIFY SEQ
SEQUENCE 2 SEQ 2
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FILE UTILITIES
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQUENCE FILE NAMING
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FILE NAME FILE0
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ERASE TITLE
NNNNN
D
NNNNN
T
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LOOP COUNTER
NNNNNNNNNNNNNN
DONE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PLOT NAME PLOTFILE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ERASE TITLE
NNNNN
NNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNN
P
LOOP COUNTER
DONE
RETURN
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNN
MENU
TRIGGER MENU
SINGLE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SA VE/RECALL
SAVE STATE
NNNNNNNNNNNNNN
COPY
PLOT
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQ
SPECIAL FUNCTIONS
DECISION MAKING
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DECR LOOP COUNTER
IF LOOP COUNTER < > 0
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQUENCE 2 SEQ 2
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQ
DONE SEQ MODIFY
This will create the following displayed lists:
INTERNAL DISK
DATA ONLY
ON
DO SEQUENCE
SEQUENCE 2

DT[LOOP]
PLOT NAME
PL[LOOP]
SINGLE
SAVE FILE 0
PLOT
DECR LOOP COUNTER
IF LOOP COUNTER 0 THEN DO
SEQUENCE 2
Sequence 1 initializes the loop counter and calls sequence 2. Sequence 2 repeats until the loop
counter reaches 0. It takes a single sweep , saves the data le and plots the display .
The data le names generated by this sequence will be:
DT000001.D1
The plot le names generated by this sequence will be:
PL00001.FP
T o run the sequence , press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PRESET
SEQUENCE 1 SEQ 1
Limit T est Example Sequence
This measurement example shows you how to create a sequence that will branch the sequence
according to the outcome of a limit test. Refer to \T esting a Device with Limit Lines ," located
earlier in this Chapter , for a procedure that shows you how to create a limit test.
F or this example , you must have already saved the following in register 1:
a series of active (visible) limit lines
an active limit test
branch to another sequence position based on the outcome of that limit test, press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQ
NEW SEQ MODIFY
SEQUENCE 1 SEQ1
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SA VE/RECALL
RECALL KEYS MENU
RECALL REG1
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQ
SPECIAL FUNCTIONS
DECISION MAKING
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
IF LIMIT TEST PASS
SEQUENCE 2 SEQ2
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
IF LIMIT TEST FAIL
SEQUENCE 3 SEQ3
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQ
DONE SEQ MODIFY
This will create a displayed list for sequence 1, as shown:

IF LIMIT TEST PASS THEN DO
SEQUENCE 2
IF LIMIT TEST FAIL THEN DO
SEQUENCE 3
test, press:
NEW SEQ MODIFY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQUENCE 2 SEQ2
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SELECT DISK
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
INTERNAL DISK
NNNNNNNNNNNNNNNNNNNN
RETURN
SA VE/RECALL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DEFINE DISK-SAVE
DATA ARRAY ON
RETURN
SAVE STATE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQ
DONE SEQ MODIFY
This will create a displayed list for sequence 2, as shown:
DATA ARRAY
ON
FILENAME
FILE 0
SAVE FILE
calls sequence 1 to retest the device , press:
NEW SEQ/MODIFY SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQUENCE 3 SEQ3
SEQ
NNNNNNNNNNNNNN
MORE
NNNNNNNNNNNNNNNNN
TITLE
DISPLA Y
NNNNN
T
NNNNN
U
NNNNN
N
NNNNN
E
NNNNN
D
NNNNN
E
NNNNN
NNNNN
NNNNN
NNNNN
E
NNNNNNNNNNNNNN
DONE
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SPECIAL FUNCTIONS
NNNNNNNNNNNNNNNNN
PAUSE
NNNNNNNNNNNNNNNNNNNN
RETURN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DO SEQUENCE
SEQUENCE 1 SEQ1
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQ
DONE SEQ MODIFY
This will create a displayed list for sequence 3, as shown:
TUNE DEVICE
SEQUENCE
PAUSE
DO SEQUENCE
SEQUENCE 1

The analyzer has the unique capability of measuring swept second and third harmonics as a
function of frequency in a real-time manner . Figure 2-47 displays the absolute power of the
fundamental and second harmonic in dBm. Figure 2-48 shows the second harmonic's power
level relative to the fundamental power in dBc . F ollow the steps listed below to perform these
measurements .
MEAS
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
INPUT PORTS
NNNNN
B to measure the power for
the fundamental frequencies .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNN
MEAS
INPUT PORTS
B to measure the power for the harmonic frequencies .
COUPLED CH OFF . Uncoupling the channels allows you to have the
separate sweeps necessary for measuring the fundamental and harmonic frequencies .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CHAN POWER [COUPLED] . Coupling the channel power allows you
to maintain the same fundamental frequency power level for both channels .
MENU
POWER and set the power level for both channels .
DISPLA Y
DUAL CHAN ON .
and position marker to desired frequency .

Figure 2-47. Fundamental and 2nd Harmonic P ower Levels in dBm
HARMONIC MEAS
NNNNNNNNNNNNNNNNNNNN
SECOND . Y ou can view both the fundamental power and
harmonic power levels at the same time . (Refer to Figure 2-47.)
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CHAN 2
DISPLA Y
D2/D1 toD2 ON . This display mode lets you see
the relationship between the fundamental and second or third harmonic in dBc . (Refer to
Figure 2-48.)


Figure 2-48. 2nd Harmonic P ower Level in dBc
Making Measurements 2-67

Measuring a Device in the Time Domain (Option 010 Only)
The HP 8753D Option 010 allows you to measure the time domain response of a device . Time
domain analysis is useful for isolating a device problem in time or in distance . Time and
distance are related by the velocity factor of your device under test. The analyzer measures
the frequency response of your device and uses an inverse F ourier transform to convert the
data to the time domain.
Time domain analysis allows you to mathematically remove individual parts of the time domain
response to see the e ect of potential design changes . Y ou can accomplish this by \gating" out
the undesirable responses .
This section shows you how to use the time domain function to measure a device response by
the following measurement examples:
transmission measurement of RF crosstalk and multi-path signal through a surface acoustic
wave (SA W) lter
re ection measurement that locates re ections along a terminated transmission line
Transmission Response in Time Domain
In this example measurement there are three components of the transmission response:
RF leakage at near zero time
the main travel path through the device (1.6 s travel time)
the \triple travel" path (4.8 s travel time)
This example procedure also shows you how time domain analysis allows you to mathematically
remove individual parts of the time domain response to see the e ect of potential design
changes . This is accomplished by \gating" out the undesirable responses . With the \gating"
capability , the analyzer time domain allows you perform \what if" analysis by mathematically
removing selected re ections and seeing the e ect in the frequency domain.

Figure 2-49. Device Connections for Time Domain Transmission Example Measurement
Gating · 2-68 Making Measurements

N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MEAS
Trans:FWD S21 (B/R)
M/
STOP
M/
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AUTO SCALE
SCALE REF
Refer to \Calibrating the Analyzer ," located at the beginning of this Chapter , for a detailed
procedure .
from 0 s to 6 s , press:
NNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SYSTEM
TRANSFORM MENU
BANDPASS
TRANSFORM ON
ST ART
STOP
M/
The other time domain modes , low pass step and low pass impulse , are described in
Chapter 6, \Application and Operation Concepts ."
REFERENCE VALUE and turn the front panel knob or enter a value from the
front panel keypad.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER F CTN
MKR SEARCH
SEARCH: MAX
The three responses shown in Figure 2-50 are the RF leakage near zero seconds , the
main travel path through the lter , and the triple travel path through the lter . Only the
combination of these responses was evident to you in the frequency domain.

Figure 2-50. Time Domain Transmission Example Measurement

TRANSFORM MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SPECIFY GATE
NNNNNNNNNNNNNNNNNNNN
CENTER
SYSTEM
1.6
, or turn the front panel knob to position the \>" center gate marker .
SPAN
1.2
or turn the front panel knob to position the \ ag" gate markers .
GATE ON
As shown in Figure 2-51, only response from the main path is displayed.
Y ou may remove the displayed response from inside the gate markers by
NNNNNNNNNNNNN
SPAN and turning the front panel knob to exchange the \ ag" marker
positions .

Figure 2-51.
Gating in a Time Domain Transmission Example Measurement
T o adjust the gate shape for the best possible time domain response , press
select between minimum, normal, wide , and maximum. Each gate has a di erent passband
atness , cuto rate , and sidelobe levels .

T able 2-2. Gate Characteristics
Gate Span Minimum 0.1 dB 48 dB 1.4/Freq Span 2.8/Freq Span
Normal 0.1 dB 68 dB 2.8/Freq Span 5.6/Freq Span
Wide 0.1 dB 57 dB 4.4/Freq Span 8.8/Freq Span
Maximum 0.01 dB 70 dB 12.7/Freq Span 25.4/Freq Span
The passband ripple and sidelobe levels are descriptive of the gate shape . The cuto time is the
time between the stop time (6 dB on the lter skirt) and the peak of the rst sidelobe , and
is equal on the left and right side skirts of the lter . Because the minimum gate span has no
passband, it is just twice the cuto time .

Figure 2-52. Gate Shape
TRANSFORM MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TRANSFORM OFF
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SCALE REF
AUTO SCALE
NNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DISPLA Y
DATA ! MEM
DISPLAY: DATA AND MEMORY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNN
SYSTEM
TRANSFORM MENU
SPECIFY GATE
GATE OFF
This places the gated response in memory . Figure 2-53 shows the e ect of removing the RF
leakage and the triple travel signal path using gating. By transforming back to the frequency
domain we see that this design change would yield better out-of-band rejection.


Figure 2-53.
Gating E ects in a Frequency Domain Example Measurement
2-72 Making Measurements

Re ection Response in Time Domain
The time domain response of a re ection measurement is often compared with the time
domain re ectometry (TDR) measurements . Like the TDR, the analyzer measures the size of
the re ections versus time (or distance). Unlike the TDR, the time domain capability of the
analyzer allows you to choose the frequency range over which you would like to make the
measurement.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MEAS
M/
STOP
1-port correction on PORT 1. Refer to Chapter 5, \Optimizing Measurement
Results ," for a detailed procedure .

Figure 2-54. Device Connections for Re ection Time Domain Example Measurement

AUTO SCALE
SCALE REF
Figure 2-55 shows the frequency domain re ection response of the cables under test. The
complex ripple pattern is caused by re ections from the adapters interacting with each
other . By transforming this data to the time domain, you can determine the magnitude of
the re ections versus distance along the cable .

Figure 2-55. Device Response in the Frequency Domain
NNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SYSTEM
TRANSFORM MENU
BANDPASS
TRANSFORM ON
NNNNNNNNNNNNNNNNNNNNNNN
LIN MAG
ST ART
STOP
The stop time corresponds to the length of the cable under test. The energy travels about
1 foot per nanosecond, or 0.3 meter/ns , in free space . Most cables have a relative velocity of
about 0.66 the speed in free space . Calculate about 3 ns/foot, or 10 ns/meter , for the stop
time when you are measuring the return trip distance to the cable end.

MORE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
VELOCITY FACTOR
CAL
and enter a velocity factor for your cable under test.
Most cables have a relative velocity of 0.66 (for polyethylene dielectrics) or 0.7
(for te on dielectrics). If you would like the markers to read actual one-way
distance rather than return trip distance , enter one-half the actual velocity
factor . Then the markers will read the actual round trip distance to the
re ection of interest rather than the \electrical length" that assumes a relative
velocity of 1.
V el ocity F actor =
is the relative permittivity of the cable dielectric .
and turn the front panel knob or enter a value from the front panel keypad.
In this example , the velocity factor was set to one-half the actual value , so the marker reads
the time and distance to the re ection.
MARKER 2
NNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER 3
NNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER 4 , turning the front panel knob or entering a value from the
front panel keypad after each key press .

Figure 2-56. Device Response in the Time Domain

The capability of making non-coaxial measurements is available with the HP 8753 family of
analyzers with TRL* (thru-re ect-line) or LRM* (line-re ect-match) calibration. F or indepth
information on TRL*/LRM* calibration, refer to Chapter 6, \Application and Operation
Concepts ."
Non-coaxial, on-wafer measurements present a unique set of challenges for error correction in
the analyzer:
The close spacing between the microwave probes makes it dicult to maintain a high degree
of isolation between the input and the output.
The type of device measured on-wafer is often not always a simple two-port.
It may be dicult to make repeatable on-wafer contacts due to the size of the device contact
pads .
Due to the simplicity of the calibration standards , TRL* or LRM* calibrations may be used
for non-coaxial applications such as on-wafer measurements . This type of calibration with
time domain gating and a variety of probe styles can provide optimal accuracy in on-wafer
measurements . At frequencies where on-wafer calibration standards are available , SOL T
calibrations can also be done and may be preferred due to the better accuracy of the SOL T
calibration method.
F or information on how to perform TRL* or LRM* calibrations , refer to the section \TRL* and
TRM* Error-Correction" in Chapter 5, \Optimizing Measurement Results ."
Non-coaxial Measurements · 2-76 Making Measurements

This chapter contains information and example procedures on the following topics:
Minimizing Source and Load Mismatches
Reducing the E ect of Spurious Responses
Eliminating Unwanted Mixing and Leakage Signals
How RF and IF Are De ned
Frequency O set Mode Operation
Di erences Between Internal and External R-Channel Inputs
P ower Meter Calibration
Conversion loss using the frequency o set mode
High dynamic range swept RF/IF conversion loss
Conversion compression using the frequency o set mode
LO to RF isolation
Where to Look for More Information
A dditional information about many of the topics discussed in this chapter is located in the
following areas:
Chapter 2, \Making Measurements ," contains step-by-step procedures for making
measurements or using particular functions .
Chapter 4, \Printing, Plotting, and Saving Measurement Results ," contains instructions
for saving to disk or the analyzer internal memory , and printing and plotting displayed
measurements .
Chapter 5, \Optimizing Measurement Results ," describes techniques and functions for
achieving the best measurement results .
Chapter 6, \Application and Operation Concepts ," contains explanatory-style information
about many applications and analyzer operation.
Making Mixer Measurements 3-1

T o ensure successful mixer measurements , the following measurement challenges must be taken
into consideration:
Minimizing Source and Load Mismatches
Reducing the E ect of Spurious Responses
Eliminating Unwanted Mixing and Leakage Signals
How RF and IF Are De ned
Frequency O set Mode Operation
Di erences Between Internal and External R-Channel Inputs
P ower Meter Calibration
Minimizing Source and Load Mismatches
When characterizing linear devices , you can use vector accuracy enhancement to
mathematically remove all systematic errors , including source and load mismatches , from your
measurement. This is not possible when the device you are characterizing is a mixer operating
over multiple frequency ranges . Therefore , source and load mismatches are not corrected for
and will add to overall measurement uncertainty .
Y ou should place attenuators at all of the test ports to reduce the measurement errors
associated with the interaction between mixer port matches and system port matches . T o avoid
overdriving the receiver , you should give extra care to selecting the attenuator located at the
mixer's IF port. F or best results , you should choose the attenuator value so that the power
incident on the analyzer R-channel input is less than 10 dBm and greater than 35 dBm.
Reducing the E ect of Spurious Responses
By choosing test frequencies (frequency list mode), you can reduce the e ect of spurious
responses on measurements by avoiding frequencies that produce IF signal path distortion.
Eliminating Unwanted Mixing and Leakage Signals
By placing lters between the mixer's IF port and the receiver's input port, you can eliminate
unwanted mixing and leakage signals from entering the analyzer's receiver . Filtering is
required in both xed and broadband measurements . Therefore , when con guring broad-band
(swept) measurements you may need to trade some measurement bandwidth for the ability to
more selectively lter signals entering the analyzer receiver .
How RF and IF Are De ned
In standard mixer measurements , the input of the mixer is always connected to the analyzer's
RF source , and the output of the mixer always produces the IF frequencies that are received by
the analyzer's receiver .
However , the ports labeled RF and IF on most mixers are not consistently connected to the
analyzer's source and receiver ports , respectively . These mixer ports are switched, depending
on whether a down converter or an up converter measurement is being performed.
It is important to keep in mind that in the setup diagrams of the frequency o set mode , the
analyzer's source and receiver ports are labeled according to the mixer port that they are
connected to .
3-2 Making Mixer Measurements
Measurement Considerations · Mixer Considerations · Analyzer Operation

In a down converter measurement where the
DOWN CONVERTER softkey is selected, the
labeled RF , connecting to the mixer RF port, and the analyzer's receiver frequency is labeled
IF , connecting to the mixer IF port.
Because the RF frequency can be greater or less than the set LO frequency in this type of
NNNNNNNNNNNNNNNNNNNNNNN
measurement, you can select either
RF > LO or
RF < LO .

Figure 3-1. Down Converter P ort Connections
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
In an up converter measurement where the
UP CONVERTER softkey is selected, the notation
on the setup diagram indicates that the analyzer's source frequency is labeled IF , connecting
to the mixer IF port, and the analyzer's receiver frequency is labeled RF , connecting to the
mixer RF port.
Because the RF frequency will always be greater than the set LO frequency in this type of
measurement, you must select only
RF > LO .

Figure 3-2. Up Converter P ort Connections
Making Mixer Measurements 3-3

Frequency O set Mode Operation
Frequency o set measurements do not begin until all of the frequency o set mode parameters
are set. These include the following:
Start and Stop IF Frequencies
Up Converter / Down Converter
The LO frequency for frequency o set mode must be set to the same value as the external LO
source . The o set frequency between the analyzer source and receiver will be set to this value .
When frequency o set mode operation begins , the receiver locks onto the entered IF signal
frequencies and then o sets the source frequency required to produce the IF . Therefore , since
it is the analyzer receiver that controls the source , it is only necessary to set the start and stop
frequencies from the receiver .
Di erences Between Internal and External R-Channel Inputs
Due to internal losses in the analyzer's test set, the power measured internally at the
R-Channel is 16 dB lower than that of the source . T o compensate for these losses , the traces
associated with the R-Channel have been o set 16 dB higher . As a result, power measured
directly at the R-Channel via the R CHANNEL IN port will appear to be 16 dB higher than its
actual value . If power meter calibration is not used, this o set in power must be accounted for
with a receiver calibration before performing measurements .
The following steps can be performed to observe this o set in power:
NNNNNNNNNNNNNNNNN
POWER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PWR RANGE MAN
Setting the power range to manual prevents the internal source attenuator from switching
when changing power levels . If you choose a di erent power range , the R-channel o set
compensation and R-channel measurement changes by the amount of the attenuator setting.
Figure 3-3.
T o prevent connector damage , use an adapter (HP part number 1250-1462) as a
connector saver for R CHANNEL IN.

Figure 3-3. R-Channel External Connection
3-4 Making Mixer Measurements

N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
INSTRUMENT MODE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FREQ OFFS MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FREQ OFFS ON
Since the LO (o set) frequency is still set to the default value of 0 Hz, the analyzer will
operate normally .
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
INPUT PORTS
NNNNN
R
Observe the 13 to 16 dB o set in measured power . The actual input power level to the
R-channel input must be 0 dBm or less , 10 dBm typical, to avoid receiver saturation
e ects . The minimum signal level must be greater than 35 dBm to provide sucient signal
for operation of the phaselock loop .
P erform a receiver calibration to remove any power o sets by pressing:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RECEIVER CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TAKE RCVR CAL SWEEP
Once completed, the R-channel should display 0 dBm. Changing power ranges will require a
recalibration of the R-channel.
Making Mixer Measurements 3-5

P ower Meter Calibration
Mixer transmission measurements are generally con gured as follows:
measured output power (W atts) / set input power (W atts)
measured output power (dBm) set input power (dBm)
F or this reason, the set input power must be accurately controlled in order to ensure
measurement accuracy .
The amplitude variation of the analyzer is speci ed 1 dB over any given source frequency .
This may give a maximum 2 dB error for a mixer transmission test setup: 1 dB for the source
over the IF range during measurement and 1 dB over the RF range during measurement.
Higher measurement accuracy may be obtained through the use of power meter calibration.
Y ou can use power meter calibration to correct for power o sets , losses , and atness variations
occurring between the analyzer source and the input to the mixer under test.
3-6 Making Mixer Measurements

Conversion Loss Using the Frequency O set Mode
Conversion loss is the measure of eciency of a mixer . It is the ratio of side-band IF power
to RF signal power , and is usually expressed in dB . (Express ratio values in dB amounts to a
subtraction of the dB power in the denominator from the dB power in the numerator .) The
mixer translates the incoming signal, (RF), to a replica, (IF), displaced in frequency by the
local oscillator , (LO). Frequency translation is characterized by a loss in signal amplitude and
the generation of additional sidebands . F or a given translation, two equal output signals are
expected, a lower sideband and an upper sideband.

Figure 3-4.
An Example Spectrum of RF , LO , and IF Signals Present in a
The analyzer allows you to make a swept RF/IF conversion loss measurement holding the LO
frequency xed. Y ou can make this measurement by using the analyzer's frequency o set
measurement mode . This mode of operation allows you to o set the analyzer's source by a
xed value , above or below the analyzer's receiver . That is , this allows you to use a device
input frequency range that is di erent from the receiver input frequency range .
The following procedure describes the swept IF frequency conversion loss measurement of a
broadband component mixer:
CW frequency = 1000 MHz
P ower = 13 dBm
R-channel input. Press:
NNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PWR RANGE MAN
POWER
The low pass lter is required to limit the range of frequencies passed into the R-channel
input port. The lter is selected to pass the IF frequencies for the measurement but
prevent the LO feedthrough and unwanted mixer products from confusing the phase lock
loop operation.
A pad is used to isolate the lter and improve the IF port match for the mixer .
The attenuation of the power splitter is used to improve the RF port match for the mixer .
Making Mixer Measurements 3-7

T o prevent connector damage , use an adapter (HP part number 1250-1462) as a
connector saver for R CHANNEL IN.

Figure 3-5. Connections for R Channel and Source Calibration
source output power , by pressing:
INSTRUMENT MODE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FREQ OFFS MENU
SYSTEM
ST ART
M/
STOP
M/
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FREQ OFFS ON
MENU
NNNNNNNNNNNNNNNNN
POWER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNN
MEAS
INPUT PORTS
R
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SYSTEM CONTROLLER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SET ADDRESSES
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ADDRESS: P MTR/HPIB
Select the appropriate power meter by pressing
POWER MTR [ ] until the correct model
number is displayed (HP 436A or HP 438A/437).
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL
PWRMTR CAL
LOSS/SENSR LISTS
CAL FACTOR SENSOR A and enter the
correction factors as listed on the power sensor . Press
ADD
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FREQUENCY
M/
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNN
NNNNNNNNNNNNNN
CAL FACTOR
DONE for each correction factor . When nished, press
DONE .
3-8 Making Mixer Measurements

press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PWRMTR CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ONE SWEEP
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TAKE CAL SWEEP
Because power meter calibration requires a longer sweep time , you may want
to reduce the number of points before pressing
TAKE CAL SWEEP . After the
power meter calibration is nished, return the number of points to its original
value and the analyzer will automatically interpolate this calibration.
CAL
RECEIVER CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TAKE RCVR CAL SWEEP
Once completed, the display should read 0 dBm.
over the RF range .

Figure 3-6.
Connections for a One-Sweep P ower Meter Calibration for Mixer Measurements
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
INSTRUMENT MODE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FREQ OFFS MENU
NNNNNNNNNNNNNNNNNNNNNNN
LO MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FREQUENCY:CW
M/
Making Mixer Measurements 3-9

RETURN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DOWN CONVERTER
NNNNNNNNNNNNNNNNN
RF<LO
actually sweeping backwards , as shown in Figure 3-7. The measurements setup diagram is
shown in Figure 3-8 .

Figure 3-7. Diagram of Measurement Frequencies

Figure 3-8. Measurement Setup from Display
VIEW MEASURE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ONE SWEEP
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL
PWRMTR CAL
TAKE CAL SWEEP
Do not reduce the number of points to perform this power meter calibration.
Reducing the number of points will turn o the receiver calibration.
The analyzer is now displaying the conversion loss of the mixer calibrated with power
meter accuracy .
3-10 Making Mixer Measurements

NNNNNNNNNNNNNNNNNNNNNNNNNNNN
SCALE REF
AUTOSCALE

Figure 3-9. Conversion Loss Example Measurement
C onv er sion l oss=g ain = output pow er input pow er
In this measurement, you set the input power and measured the output power . Figure 3-9
shows the absolute loss through the mixer versus mixer output frequency . If the mixer
under test contained built-in ampli cation, then the measurement results would have shown
conversion gain.
Making Mixer Measurements 3-11
·

High Dynamic Range Swept RF/IF Conversion Loss
The HP 8753D's frequency o set mode enables the testing of high dynamic range frequency
converters (mixers), by tuning the analyzer's high dynamic range receiver above or below its
source , by a xed o set. This capability allows the complete measurement of both pass and
reject band mixer characteristics .
The analyzer has a 35 dB dynamic range limitation on measurements made directly with its R
(phaselock) channel. F or this reason, the measurement of high dynamic range mixing devices
(such as mixers with built in ampli cation and ltering) with greater than 35 dB dynamic range
must be made on either the analyzer's A or B channel, with a reference mixer providing input
to the analyzer's R-channel for phaselock.
This example describes the swept IF conversion loss measurement of a mixer and lter . The
output ltering demonstrates the analyzer's ability to make high dynamic range measurements .
T o avoid the complexity of performing a separate power meter calibration over the RF
frequency range while the mixer under test and reference mixer are operating, a broad band
power meter calibration is used. The broad band calibration covers the entire range of IF and
RF frequencies .
ST ART
M/
STOP
2.5
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PWR RANGE MAN
MENU
POWER
T o prevent connector damage , use an adapter (HP part number 1250-1462) as a
connector saver for R CHANNEL IN.
3-12 Making Mixer Measurements


Figure 3-10. Connections for Broad Band P ower Meter Calibration
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SYSTEM CONTROLLER
SET ADDRESSES
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ADDRESS: P MTR/HPIB
Select the appropriate power meter by pressing
POWER MTR [ ] until the correct model
number is displayed (HP 436A or HP 438A/437).
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL
PWRMTR CAL
LOSS/SENSR LISTS
CAL FACTOR SENSOR A and enter the
correction factors as listed on the power sensor . Press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FREQUENCY
ADD
M/
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL FACTOR
NNNNNNNNNNNNNN
DONE for each correction factor . When nished, press
DONE
NNNNNNNNNNNNNNNNNNNN
RETURN .
NNNNN
MEAS
INPUT PORTS
B
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL
PWRMTR CAL
ONE SWEEP
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TAKE CAL SWEEP
Making Mixer Measurements 3-13

Because power meter calibration requires a longer sweep time , you may want
to reduce the number of points before pressing
TAKE CAL SWEEP . After the
power meter calibration is nished, return the number of points to its original
value and the analyzer will automatically interpolate this calibration.

Figure 3-11. Connections for Receiver Calibration
M/
STOP
RECEIVER CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TAKE RCVR CAL SWEEP
CAL
Once completed, the analyzer should display 0 dBm.
values are as follows:
CW frequency = 1500 MHz
source power = 13 dBm
3-14 Making Mixer Measurements


Figure 3-12.
Connections for a High Dynamic Range Swept IF Conversion Loss Measurement
Making Mixer Measurements 3-15

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
INSTRUMENT MODE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FREQ OFFS MENU
NNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LO MENU
FREQUENCY:CW
M/
RETURN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DOWN CONVERTER
NNNNNNNNNNNNNNNNN
RF>LO
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FREQ OFFS ON
In this low-side LO , down converter measurement, the analyzer's source frequency range
will be o set higher than the receiver frequency range . The source frequency range can be
determined from the following equation:
receiver frequency range (100 { 1000 MHz) + LO frequency (1500 MHz) = 1.6-2.5 GHz
VIEW MEASURE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AUTOSCALE
SCALE REF
Figure 3-13 shows the conversion loss of this low-side LO , mixer with output ltering.
dynamic range limit of the R Channel. If the mixer under test also contained ampli cation,
then this dynamic range would have been even greater due to the conversion gain of the
mixer .

Figure 3-13. Example of Swept IF Conversion Loss Measurement
3-16 Making Mixer Measurements

Fixed IF Mixer Measurements
A xed IF can be produced by using both a swept RF and LO that are o set by a certain
frequency . With proper ltering, only this o set frequency will be present at the IF port of the
mixer .
This measurement requires two external RF sources as stimuli. Figure 3-15 shows the hardware
con guration for the xed IF conversion loss measurement. This example measurement
procedure uses the analyzer's test sequence function for automatically controlling the two
external synthesizers (with SCPI commands), and making a conversion loss measurement in
tuned receiver mode . The test sequence function is an instrument automation feature internal
to the analyzer . F or more information on the test sequence function refer to \T est Sequencing"
located in Chapter 2.
The analyzer's tuned receiver mode allows you to tune its receiver to an arbitrary frequency
and measure signal power . This is only possible if the signal you wish to analyze is at an exact
known frequency . Therefore , the RF and LO sources must be synthesized and synchronized
with the analyzer's time base .
The following sequence initializes and calibrates the network analyzer . It then initializes the
two external sources prior to measurement. This sequence includes:
putting the network analyzer into tuned receiver mode
setting up a frequency list sweep of 26 points
performing a response calibration
prompting the user to connect a mixer to the test set up
initializing a loop counter value to 26
addressing and con guring the two sources
calling the next measurement sequence
external RF source to 19 and the external LO source to 21.
programming language and that their output power is switched on.
Y ou may have to consult the User's Guide of the external source being used to
determine how to set the source to receive SCPI commands .
connector on the rear panel of the analyzer (a BNC tee must be used).
If the 10 MHz reference signals of the external sources are connected together ,
then it will only be necessary to connect one reference signal from one of the
sources to the EXT REF connector of the analyzer .
Making Mixer Measurements 3-17
Tuned Receiver Mode · Sequence 1 Setup


Figure 3-14. Connections for a Response Calibration
T o enter the following sequence commands that require titling, an external
keyboard may be used for convenience .
NEW SEQ/MODIFY SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQUENCE 1 SEQ1
SEQ
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SA VE/RECALL
SELECT DISK
INTERNAL MEMORY
NNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RETURN (Select the preset state .)
Putting the Analyzer into Tuned Receiver Mode
LOCAL
SYSTEM CONTROLLER
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SYSTEM
INSTRUMENT MODE
TUNED RECEIVER
Setting Up a Frequency List Sweep of 26 P oints
SWEEP TYPE MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
EDIT LIST
NNNNNNNNNNN
ADD
MENU
NNNNNNNNNNNNNNNNNNNNNNN
CW FREQ
M/
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NUMBER OF POINTS
NNNNNNNNNNNNNN
DONE
NNNNNNNNNNNNNN
DONE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LIST FREQ
P erforming a Response Calibration
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNN
MEAS
INPUT PORTS
B
NNNNNNNNNNNNNN
MORE
NNNNNNNNNNNNNNNNN
TITLE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ERASE TITLE
DISPLA Y
3-18 Making Mixer Measurements

NNNNNNNNNNNNNN
DONE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQ
SPECIAL FUNCTIONS
PERIPHERAL HPIB ADDR
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TITLE TO PERIPHERAL
NNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DISPLA Y
MORE
TITLE
ERASE TITLE
FREQ:MODE CW;CW 100MHZ
NNNNNNNNNNNNNN
DONE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SPECIAL FUNCTIONS
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PERIPHERAL HPIB ADDR
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TITLE TO PERIPHERAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CALIBRATE MENU
NNNNNNNNNNNNNNNNNNNNNNNNNN
RESPONSE
CAL
NNNNNNNNNNNNNN
THRU
Prompting the User to Connect a Mixer to the T est Set Up
NNNNNNNNNNNNN
NNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DISPLA Y
MORE
TITLE
ERASE TITLE
CONNECT MIXER
NNNNNNNNNNNNNN
DONE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQ
SPECIAL FUNCTIONS
NNNNNNNNNNNNNNNNN
PAUSE
Initializing a Loop Counter V alue to 26
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQ
SPECIAL FUNCTIONS
DECISION MAKING
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LOOP COUNTER
SCALE REF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
REFERENCE POSITION
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
REFERENCE VALUE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MENU
TRIGGER MENU
MANUAL TRG ON POINT
A ddressing and Con guring the Two Sources
NNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DISPLA Y
MORE
TITLE
ERASE TITLE
FREQ:MODE CW;CW 500MHZ;:FREQ:CW:STEP 100MHZ
NNNNNNNNNNNNNN
DONE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQ
SPECIAL FUNCTIONS
PERIPHERAL HPIB ADDR
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TITLE TO PERIPHERAL
NNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DISPLA Y
MORE
TITLE
ERASE TITLE
POW:LEV 13DBM
NNNNNNNNNNNNNN
DONE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQ
SPECIAL FUNCTIONS
PERIPHERAL HPIB ADDR
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TITLE TO PERIPHERAL
NNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DISPLA Y
MORE
TITLE
ERASE TITLE
FREQ:MODE CW;CW 600MHZ;:FREQ:CW:STEP 100MHZ
NNNNNNNNNNNNNN
DONE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PERIPHERAL HPIB ADDR
SEQ
SPECIAL FUNCTIONS
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TITLE TO PERIPHERAL
Making Mixer Measurements 3-19

Calling the Next Measurement Sequence
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQ
DO SEQUENCE
SEQUENCE 2 SEQ2
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQ
DONE SEQ MODIFY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQ
NEW SEQ/MODIFY SEQ
SEQUENCE 1 SEQ 1 and the analyzer will display the
following sequence commands:
RECALL PRST STATE
SYSTEM CONTROLLER
TUNED RECEIVER
EDIT LIST
ADD
CW FREQ
NUMBER OF POINTS
DONE
DONE
LIST FREQ
B
TITLE
POW:LEV 6DBM
PERIPHERAL HPIB ADDR
TITLE TO PERIPHERAL
TITLE
FREQ:MODE CW;CW 100MHZ
TITLE TO PERIPHERAL
CALIBRATE: RESPONSE
CAL STANDARD
DONE CAL CLASS
TITLE
CONNECT MIXER
PAUSE
LOOP COUNTER
SCALE/DIV
REFERENCE POSITION
REFERENCE VALUE
MANUAL TRG ON POINT
TITLE
FREQ:MODE CW;CW 500MHZ;:FREQ:CW:STEP 100MHZ
TITLE TO PERIPHERAL
TITLE
POW:LEV 13DBM
PERIPHERAL HPIB ADDR
TITLE TO PERIPHERAL
TITLE
3-20 Making Mixer Measurements

TITLE TO PERIPHERAL
DO SEQUENCE
SEQUENCE 2
The following sequence makes a series of measurements until all 26 CW measurements are
made and the loop counter value is equal to zero . This sequence includes:
incrementing the source frequencies
decrementing the loop counter
T o enter the following sequence commands that require titling, an external
keyboard may be used for convenience .
SEQ
DONE SEQ MODIFY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQ
NEW SEQ/MODIFY SEQ
SEQUENCE 2 SEQ2
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SPECIAL FUNCTIONS
NNNNNNNNNNNNNNNNNNNN
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TRIGGER MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MANUAL TRG ON POINT
MENU
Incrementing the Source Frequencies
NNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DISPLA Y
MORE
TITLE
ERASE TITLE
FREQ:CW UP
NNNNNNNNNNNNNN
DONE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SPECIAL FUNCTIONS
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PERIPHERAL HPIB ADDR
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TITLE TO PERIPHERAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PERIPHERAL HPIB ADDR
TITLE TO PERIPHERAL
Decrementing the Loop Counter
DECISION MAKING
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DECR LOOP COUNTER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
IF LOOP COUNTER<>0
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQUENCE 2 SEQ2
NNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DISPLA Y
MORE
TITLE
ERASE TITLE
MEASUREMENT COMPLETED
NNNNNNNNNNNNN
N
DONE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQ
DONE SEQ MODIFY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQ
NEW SEQ/MODIFY SEQ
SEQUENCE 2 SEQ 2 and the analyzer will display the
following sequence commands:
Making Mixer Measurements 3-21

TITLE
FREQ:CW UP
PERIPHERAL HPIB ADDR
TITLE TO PERIPHERAL
PERIPHERAL HPIB ADDR
TITLE TO PERIPHERAL
DECR LOOP COUNTER
IF LOOP COUNTER <>0 THEN DO
SEQUENCE 2
TITLE
MEASUREMENT COMPLETED
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQ
DONE SEQ MODIFY
DO SEQUENCE
SEQUENCE2 SEQ2
When the prompt CONNECT MIXER appears , connect the equipment as shown in Figure 3-15.

Figure 3-15. Connections for a Conversion Loss Using the Tuned Receiver Mode
When the sequences are nished you should have a result as shown in Figure 3-16.
3-22 Making Mixer Measurements
·


Figure 3-16. Example Fixed IF Mixer Measurement
The displayed trace represents the conversion loss of the mixer at 26 points . Each point
corresponds to one of the 26 di erent sets of RF and LO frequencies that were used to create
the same xed IF frequency .
Making Mixer Measurements 3-23

Phase or Group Delay Measurements
F or information on group delay principles , refer to \Group Delay Principles" in Chapter 6.
The accuracy of this measurement depends on the quality of the mixer that is being used for
calibration and how well this mixer has been characterized. The following measurement must
be performed with a broadband calibration mixer that has a known group delay . The following
table lists the speci cations of two mixers that may be used for calibration:
.03 to 3 GHz .5 ns
dc to 1250 MHz .6 ns
source is set to the following values:
CW frequency = 1000 MHz
power = 13 dBm
.
output power by pressing:
M/
SP AN
M/
NNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MENU
POWER
PWR RANGE MAN
(ZFM-4) between PORT 1 and PORT 2.
T o prevent connector damage , use an adapter (HP part number 1250-1462) as a
connector saver for R CHANNEL IN.
3-24 Making Mixer Measurements


Figure 3-17. Connections for a Group Delay Measurement
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
INSTRUMENT MODE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FREQ OFFS MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
VIEW MEASURE
NNNNNNNNNNNNNNNNNNNNNNN
LO MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FREQUENCY:CW
M/
NNNNNNNNNNNNNNNNNNN
RETURN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DOWN CONVERTER
NNNNNNNNNNNNNNNNN
RF<LO
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FREQ OFFS ON
F ORMA T
DELAY
Making Mixer Measurements 3-25

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
INPUT PORTS
NNNNNNNNNNN
B/R
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNN
CAL
CALIBRATE MENU
RESPONSE
THRU
the delay equal to the \calibration" mixer delay (for example 0.6 ns) by pressing:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ELECTRICAL DELAY
0.6
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AUTOSCALE

Figure 3-18. Group Delay Measurement Example
The displayed measurement trace shows the device under test delay , relative to the
\calibration" mixer . This measurement is also useful when the device under test has built-in
ltering, which requires >30 dB range (the maximum of R input). PORT 1 to PORT 2 range is
>100 dB .
3-26 Making Mixer Measurements

Amplitude and Phase Tracking
Using the same measurement set-up as in \Phase or Group Delay Measurements ," you can
determine how well two mixers track each other in terms of amplitude and phase .
following exception:
In step 7, select
PHASE .
Once the analyzer has displayed the measurement results , press
DATA !MEM .
DISPLA Y
NNNNNNNNNNNNNNNNNNNNNNNNN
DATA/MEM .
The resulting trace should represent the amplitude and phase tracking of the two mixers .
Making Mixer Measurements 3-27
· · Press

Conversion Compression Using the Frequency O set Mode
Conversion compression is a measure of the maximum RF input signal level, where the mixer
provides linear operation. The conversion loss is the ratio of the IF output level to the RF input
level. This value remains constant over a speci ed input power range . When the input power
level exceeds a certain maximum, the constant ratio between IF and RF power levels will begin
to change . The point at which the ratio has decreased 1 dB is called the 1 dB compression
point. See Figure 3-19.

Figure 3-19.
Conversion Loss and Output P ower as a Function of Input P ower Level Example
compression begins and the mixer saturates .
The following example uses a ratio of mixer output to input power and a marker search
function to locate a mixer's 1 dB compression point.
CW frequency = 600 MHz
P ower = 13 dBm
.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SWEEP TYPE MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
POWER SWEEP
NNNNNNNNNNNNNNNNNNNN
RETURN
NNNNNNNNNNNNNNNNNNNNNNN
CW FREQ
M/
NNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
POWER
PWR RANGE MAN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNN
POWER RANGES
RANGE 0
ST ART
STOP
3-28 Making Mixer Measurements

T o prevent connector damage , use an adapter (HP part number 1250-1462) as a
connector saver for R CHANNEL IN.

Figure 3-20. Connections for the First P ortion of Conversion Compression Measurement
NNNNN
MEAS
INPUT PORTS
R
data/memory , press:
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DATA ! MEMORY
NNNNNNNNNNNNNNNNNNNNNNNNNN
DATA/MEM
This removes the loss between the output of the mixer and the input to the receiver , and
provides a linear power sweep for use in subsequent measurements .
Making Mixer Measurements 3-29

T o prevent connector damage , use an adapter (HP part number 1250-1462) as a
connector saver for R CHANNEL IN.

Figure 3-21.
Connections for the Second P ortion of Conversion Compression Measurement
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
INSTRUMENT MODE
FREQ OFFS MENU
NNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LO MENU
FREQUENCY:CW
M/
RETURN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
UP CONVERTER
RF>LO
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FREQ OFFS ON
In this low-side LO , up converter measurement, the analyzer source frequency is o set lower
than the receiver frequency . The analyzer source frequency can be determined from the
following equation:
receiver frequency (800 MHz) LO frequency (600 MHz) = 200 MHz
3-30 Making Mixer Measurements

The measurements setup diagram is shown in Figure 3-22.

Figure 3-22. Measurement Setup Diagram Shown on Analyzer Display
VIEW MEASURE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AUTO SCALE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER F CTN
MKR SEARCH ON
SEARCH:MAX
MARKER
MKR ZERO
MARKER F CTN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MKR SEARCH ON
NNNNNNNNNNNNNNNNNNNN
TARGET
The measurement results show the mixer's 1 dB compression point. By changing the target
value , you can easily locate other compression points (for example , 0.5 dB , 3 dB). See
Figure 3-22.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER
MODE
MODE OFF .
Making Mixer Measurements 3-31


Figure 3-23.
Example Swept P ower Conversion Compression Measurement
3-32 Making Mixer Measurements
j

Isolation is the measure of signal leakage in a mixer . F eedthrough is speci cally the forward
signal leakage to the IF port. High isolation means that the amount of leakage or feedthrough
between the mixer's ports is very small. Isolation measurements do not use the frequency
o set mode . Figure 3-24 illustrates the signal ow in a mixer .

Figure 3-24. Signal Flow in a Mixer Example
The RF and LO feedthrough signals may appear at the mixer IF output, together with the
desired IF signal.
The LO to RF isolation and the LO feedthrough are typically measured with the third port
terminated in 50 ohms . Measurement of the RF feedthrough is made as the LO signal is being
applied to the mixer .
LO to RF Isolation
.
M/
STOP
M/
N
POWER
NNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PWR RANGE MAN
MENU
This source stimulates the mixer's LO port.
NNNNNNNNNNN
MEAS
INPUT PORTS
B/R
Making Mixer Measurements 3-33


Figure 3-25. Connections for a Response Calibration
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNN
P erform a response calibration by pressing
CALIBRATE MENU
RESPONSE
THRU .
A full 2 port calibration will increase the accuracy of isolation measurements .
Refer to Chapter 5, \Optimizing Measurement Results ."

Figure 3-26. Connections for a Mixer Isolation Measurement
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AUTO SCALE
The measurement results show the mixer's LO to RF isolation.
3-34 Making Mixer Measurements
· ·


Figure 3-27. Example Mixer LO to RF Isolation Measurement
The procedure and equipment con guration necessary for this measurement are very similar
to those above , with the addition of an external source to drive the mixer's LO port as we
measure the mixer's RF feedthrough. RF feedthrough measurements do not use the frequency
o set mode .
CW frequency = 300 MHz
P ower = 10 dBm
.
M/
STOP
M/
NNNNNNNNNNNNNNNNN
POWER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PWR RANGE MAN
MENU
This signal stimulates the mixer's RF port.
NNNNNNNNNNN
MEAS
INPUT PORTS
B/R
Isolation is dependent on LO power level and frequency . T o ensure good
test results , you should choose these parameters as close to actual operating
conditions as possible .
Making Mixer Measurements 3-35

Figure 3-28. Connections for a Response Calibration
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNN
P erform a response calibration by pressing
CALIBRATE MENU
RESPONSE
THRU

Figure 3-29. Connections for a Mixer RF F eedthrough Measurement
Y ou may see spurious responses on the analyzer trace due to interference
caused by LO to IF leakage in the mixer . This can be reduced with averaging or
by reducing the IF bandwidth.
3-36 Making Mixer Measurements


Figure 3-30. Example Mixer RF F eedthrough Measurement
Y ou can measure the IF to RF isolation in a similar manner , but with the following
modi cations:
use the analyzer source as the IF signal drive
view the leakage signal at the RF port
Making Mixer Measurements 3-37



Printing, Plotting, and Saving Measurement
This chapter contains instructions for the following tasks:
Printing or plotting your measurement results
Con guring a print function
De ning a print function
Printing one measurement per page
Printing multiple measurements per page
Con guring a plot function
De ning a plot function
Plotting one measurement per page using a pen plotter
Plotting multiple measurements per page using a pen plotter
Plotting a measurement to disk
Outputting plot les from a PC to a plotter
Outputting plot les from a PC to an HPGL compatible printer
Outputting single page plots using a printer
Outputting multiple plots to a single page using a printer
Plotting Multiple Measurements per page from disk
Titling the displayed measurement
Con guring the analyzer to produce a time stamp
Aborting a print or plot process
Printing or plotting the list values or operating parameters
Solving problems with printing or plotting
Saving and recalling instrument states
Saving an instrument state
Re-saving an instrument state
F ormatting a disk
Solving problems with saving or recalling les
Printing, Plotting, and Saving Measurement Results 4-1

Where to Look for More Information
A dditional information about many of the topics discussed in this chapter is located in the
following areas:
Chapter 2, \Making Measurements ," contains step-by-step procedures for making
measurements or using particular functions .
Chapter 8, \Menu Maps ," shows softkey menu relationships .
Chapter 9, \Key De nitions ," describes all the front panel keys , softkeys , and their
corresponding HP-IB commands .
Chapter 11, \Compatible P eripherals ," lists measurement and system accessories , and other
applicable equipment compatible with the analyzers . An HP-IB programming overview is also
included.
4-2 Printing, Plotting, and Saving Measurement Results

Printing or Plotting Y our Measurement Results
Y ou can print your measurement results to the following peripherals:
printers with HP-IB interfaces
printers with parallel interfaces
printers with serial interfaces
Y ou can plot your measurement results to the following peripherals:
HPGL compatible printers with HP-IB interfaces
HPGL compatible printers with parallel interfaces
plotters with HP-IB interfaces
plotters with parallel interfaces
plotters with serial interfaces
Refer to the \Compatible P eripherals" chapter for a list of recommended peripherals .
Con guring a Print Function
All copy con guration settings are stored in non-volatile memory . Therefore , they are not
a ected if you press
or switch o the analyzer power .
Printer Interface Recommended Cables
P arallel HP 92284A

Figure 4-1. Printer Connections to the Analyzer
Printing, Plotting, and Saving Measurement Results 4-3
Serial HP 24542G

SET ADDRESSES
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PRINTER PORT
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PRNTR TYPE until the correct printer choice
appears:
NNNNNNNNNNNNNNNNNNNNNNN
DeskJet (except for HP DeskJet 540 and DeskJet 850C)
NNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNN
Epson-P2 (printers that conform to the ESC/P2 printer control language)
DJ 540 (for use with the HP DeskJet 540 and DeskJet 850C)
DJ 540 converts 100 dpi raster information to 300 dpi raster format.
If your DeskJet printer does not support the 100 dpi raster format and your
printing results seem to be less than normal size , select
DJ 540 .
PRNTR PORT HPIB if your printer has an HP-IB interface , and then con gure the
print function as follows:
a. Enter the HP-IB address of the printer , followed by
.
b.
SYSTEM CONTROLLER if there is no external controller connected to
the HP-IB bus .
c. Press
USE PASS CONTROL if there is an external controller connected to
the HP-IB bus .
PARALLEL if your printer has a parallel (centronics) interface , and then con gure
the print function as follows:
and then select the parallel port interface function by pressing
LOCAL
PARALLEL
until the correct function appears .
PARALLEL [COPY] , the parallel port is dedicated for normal copy
device use (printers or plotters).
PARALLEL [GPIO] , the parallel port is dedicated for general purpose
I/O , and cannot be used for printing or plotting.
4-4 Printing, Plotting, and Saving Measurement Results

SERIAL if your printer has a serial (RS-232) interface , and then con gure the print
function as follows:
PRINTER BAUD RATE and enter the printer's baud rate , followed by
.
a. Press
b. T o select the transmission control method that is compatible with your printer , press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
XMIT CNTRL (transmit control - handshaking protocol) until the correct method
appears .
Xon-Xoff , the handshake method allows the printer to control the
data exchange by transmitting control characters to the network analyzer .
DTR-DSR , the handshake method allows the printer to control the data
exchange by setting the electrical voltage on one line of the RS-232 serial cable .
DTR-DSR handshake takes place in the hardware rather than the
rmware or software , it is the fastest transmission control method.
De ning a Print Function
The print de nition is set to default values whenever the power is cycled.
However , you can save the print de nition by saving the instrument state .
DEFINE PRINT .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PRINT: COLOR .
PRINT: MONOCHROME if you are using a black and white printer , or you want just
black and white from a color printer .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PRINT: COLOR if you are using a color printer .
AUTO-FEED until the correct choice (ON or OFF) is high-lighted.
AUTO-FEED ON if you want to print one measurement per page .
AUTO-FEED OFF if you want to print multiple measurements per page .
Laser printers and some DeskJet printers do not begin to print until a full page ,
or a partial page and a form feed, have been received.
Printing, Plotting, and Saving Measurement Results 4-5

If Y ou are Using a Color Printer
PRINT COLORS .
color .
Y ou can set all the print elements to black to create a hardcopy in black and
white .
Since the media color is white or clear , you could set a print element to white if
you do not want that element to appear on your hardcopy .
T o Reset the Printing P arameters to Default V alues
DEFINE PRINT
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DEFAULT PRNT SETUP .
T able 4-1. Default V alues for Printing P arameters
Printing P arameter Default
A uto F eed ON
Channel 1 Data Magenta
Channel 1 Memory Green
Channel 2 Data Blue
Channel 2 Memory Red
Printing One Measurement P er P age
\De ning a Print Function" located earlier in this chapter .
COPY
PRINT MONOCHROME .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
If you de ned the
AUTO-FEED OFF , press
PRINTER FORM FEED after the message COPY
OUTPUT COMPLETED appears .
4-6 Printing, Plotting, and Saving Measurement Results

Printing Multiple Measurements P er P age
\De ning a Print Function" located earlier in this chapter .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
COPY
DEFINE PRINT and then press
AUTO-FEED until the softkey label appears as
AUTO-FEED OFF .
RETURN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PRINT MONOCHROME to print a measurement on the rst half page .
example of a hardcopy where two measurements appear .
COPY
PRINT MONOCHROME to print a measurement on the second half page .
This feature will not work for all printers due to di erences in printer
resolution.

Figure 4-2. Printing Two Measurements
Printing, Plotting, and Saving Measurement Results 4-7

Con guring a Plot Function
All copy con guration settings are stored in non-volatile memory . Therefore , they are not
a ected if you press
or switch o the analyzer power .
P eripheral Interface Recommended Cables
P arallel HP 92284A
Figure 4-3. P eripheral Connections to the Analyzer
If Y ou are Plotting to an HPGL/2 Compatible Printer
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LOCAL
SET ADDRESSES
PRINTER PORT and then press
PRNTR TYPE until the correct
printer choice appears:
NNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNN
DeskJet (only DeskJet 1200C and DeskJet 1600C)
LaserJet (only LaserJet III and IV)
NNNNNNNNNNNNNNNNNNNNNNNNNN
Epson-P2 (printers that conform to the ESC/P2 printer control language)
4-8 Printing, Plotting, and Saving Measurement Results

PRNTR PORT HPIB if your printer has an HP-IB interface , and then con gure the
print function as follows:
a. Enter the HP-IB address of the printer (default is 01), followed by
.
b.
LOCAL
SYSTEM CONTROLLER if there is no external controller connected to
the HP-IB bus .
c. Press
USE PASS CONTROL if there is an external controller connected to
the HP-IB bus .
PARALLEL if your printer has a parallel (centronics) interface , and then con gure
the print function as follows:
and then select the parallel port interface function by pressing
PARALLEL
until the correct function appears .
PARALLEL [COPY] , the parallel port is dedicated for normal copy
device use (printers or plotters).
PARALLEL [GPIO] , the parallel port is dedicated for general purpose
I/O , and cannot be used for printing or plotting.
SERIAL if your printer has a serial (RS-232) interface , and then con gure the print
function as follows:
a. Press
PRINTER BAUD RATE and enter the printer's baud rate , followed by
.
b. T o select the transmission control method that is compatible with your printer , press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
XMIT CNTRL (transmit control - handshaking protocol) until the correct method
appears .
Xon-Xoff , the handshake method allows the printer to control the
data exchange by transmitting control characters to the network analyzer .
DTR-DSR , the handshake method allows the printer to control the data
exchange by setting the electrical voltage on one line of the RS-232 serial cable .
DTR-DSR handshake takes place in the hardware rather than the
rmware or software , it is the fastest transmission control method.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LOCAL
SET ADDRESSES
PLOTTER PORT and then
PLTR TYPE [HPGL PRT] appears .
Printing, Plotting, and Saving Measurement Results 4-9

If Y ou are Plotting to a P en Plotter
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PLOTTER PORT and then
LOCAL
SET ADDRESSES
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PLTR TYPE [PLOTTER] appears .
PLTR PORT HPIB if your plotter has an HP-IB interface , and then con gure the
plot function as follows:
a. Enter the HP-IB address of the printer (default is 05), followed by
.
b.
SYSTEM CONTROLLER if there is no external controller connected to
the HP-IB bus .
c. Press
USE PASS CONTROL if there is an external controller connected to
the HP-IB bus .
PARALLEL if your printer has a parallel (centronics) interface , and then con gure
the print function as follows:
LOCAL
and then select the parallel port interface function by pressing
until the correct function appears .
PARALLEL [COPY] , the parallel port is dedicated for normal copy
device use (printers or plotters).
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PARALLEL [GPIO] , the parallel port is dedicated for general purpose
I/O , and cannot be used for printing or plotting.
NNNNNNNNNNNNNNNNNNN
SERIAL if your printer has a serial (RS-232) interface , and then con gure the print
function as follows:
a. Press
PRINTER BAUD RATE and enter the printer's baud rate , followed by
.
b. T o select the transmission control method that is compatible with your printer , press
XMIT CNTRL (transmit control - handshaking protocol) until the correct method
appears .
NNNNNNNNNNNNNNNNNNNNNNNNN
Xon-Xoff , the handshake method allows the printer to control the
data exchange by transmitting control characters to the network analyzer .
DTR-DSR , the handshake method allows the printer to control the data
exchange by setting the electrical voltage on one line of the RS-232 serial cable .
NNNNNNNNNNNNNNNNNNNNNN
DTR-DSR handshake takes place in the hardware rather than the
rmware or software , it is the fastest transmission control method.
4-10 Printing, Plotting, and Saving Measurement Results

If Y ou are Plotting to a Disk Drive
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNN
DISK .
SET ADDRESSES
PLOTTER PORT
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SELECT DISK and select the disk drive that you will plot to .
N
INTERNAL DISK if you will plot to the analyzer internal disk drive .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
EXTERNAL DISK if you will plot to a disk drive that is external to the analyzer .
Then con gure the disk drive as follows:
a.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CONFIGURE EXT DISK
drive (default is 00) followed by
.
b.
DISK UNIT NUMBER and enter the drive where your disk is located,
.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
If your storage disk is partitioned, press
VOLUME NUMBER and enter the volume number
where you want to store the instrument state le .
Printing, Plotting, and Saving Measurement Results 4-11

De ning a Plot Function
The plot de nition is set to default values whenever the power is cycled.
However , you can save the plot de nition by saving the instrument state .
DEFINE PLOT .
your plot:
PLOT DATA ON if you want the measurement data trace to appear on your plot.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PLOT MEM ON if you want the displayed memory trace to appear on your plot.
PLOT GRAT ON if you want the graticule and the reference line to appear on your
plot.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PLOT TEXT ON if you want all of the displayed text to appear on your plot. (This
does not include the marker values or softkey labels .)
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PLOT MKR ON if you want the displayed markers , and marker values , to appear on
your plot.

Figure 4-4. Plot Components A vailable through De nition
Selecting A uto-F eed
AUTO-FEED until the correct choice is high-lighted.
AUTO-FEED ON if you want a \page eject" sent to the plotter or HPGL compatible
printer after each time you press
PLOT .
AUTO-FEED OFF if you want multiple plots on the same sheet of paper .
4-12 Printing, Plotting, and Saving Measurement Results

AUTO-FEED ON when you are plotting to a quadrant.
Selecting P en Numbers and Colors
MORE and select the plot element where you want to change the pen number . F or
PEN NUM DATA and then modify the pen number . The pen number selects
the color if you are plotting to an HPGL/2 compatible color printer .
after each modi cation.
The following color assignments are valid for HPGL/2 compatible color printers
only . When using word processor or graphics presentation programs , di erent
colors may be assigned to the pen numbers .
T able 4-2. Default P en Numbers and Corresponding Colors
T able 4-3. Default P en Numbers for Plot Elements
Corresponding K ey Plot Element Channel 1
FFFFFFFFFFFFFFFFFFFFFFFFFFFFFF
PEN NUM DATA Measurement Data Trace 2 3
PEN NUM MEMORY Displayed Memory Trace 5 6
PEN NUM GRATICULE Graticule and Reference Line 1 1
FFFFFFFFFFFFFFFFFFFFFFFFFFFFFF
PEN NUM TEXT Displayed T ext 7 7
PEN NUM MARKER Displayed Markers and V alues 7 7
Y ou can set all the pen numbers to black for a plot in black and white .
Y ou must de ne the pen numbers for each measurement channel (channel 1
and channel 2).
Printing, Plotting, and Saving Measurement Results 4-13

MORE and select each plot element line type that you want to modify .
LINE TYPE DATA to modify the line type for the data trace . Then enter the new
line type (see Figure 4-5), followed by
.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LINE TYPE MEMORY to modify the line type for the memory trace . Then enter the
new line type (see Figure 4-5), followed by
.
T able 4-4. Default Line Types for Plot Elements
Plot Elements Channel 1
Data Trace 7 7
Memory Trace 7 7

Figure 4-5. Line Types A vailable
Y ou must de ne the line types for each measurement channel (channel 1 and
channel 2).
4-14 Printing, Plotting, and Saving Measurement Results

SCALE PLOT until the selection appears that you want.
SCALE PLOT [FULL] if you want the normal scale selection for plotting. This
includes space for all display annotations such as marker values and stimulus values . The
entire analyzer display ts within the de ned boundaries of P1 and P2 on the plotter ,
while maintaining the exact same aspect ratio as the display .
SCALE PLOT [GRAT] if you want the outer limits of the graticule to correspond to
the de ned P1 and P2 scaling point on the plotter . (Intended for plotting on preprinted
rectangular or polar) forms .

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Figure 4-6. Locations of P1 and P2 in
SCALE PLOT [GRAT] Mode
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PLOT SPEED until the plot speed appears that you want.
PLOT SPEED [FAST] for normal plotting.
PLOT SPEED [SLOW] for plotting directly on transparencies .(The slower speed
provides a more consistent line width.)
Printing, Plotting, and Saving Measurement Results 4-15

T o Reset the Plotting P arameters to Default V alues
DEFINE PLOT
NNNNNNNNNNNNNN
MORE
NNNNNNNNNNNNNN
MORE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DEFAULT PLOT SETUP .
T able 4-5. Plotting P arameter Default V alues
Plotting P arameter Default
Select Quadrant Full page
A uto F eed ON
De ne Plot All plot elements on
Plot Speed F ast
Line Type 7 (solid line)
Plotting One Measurement P er P age Using a P en Plotter
Plot Function" located earlier in this chapter .
COPY
PLOT .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
If you de ned the
AUTO-FEED OFF , press
PLOTTER FORM FEED after the message COPY
OUTPUT COMPLETED appears .
4-16 Printing, Plotting, and Saving Measurement Results

Plotting Multiple Measurements P er P age Using a P en Plotter
Plot Function" located earlier in this chapter .
COPY
SEL QUAD .
hardcopy . The following quadrants are available:
LEFT UPPER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LEFT LOWER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RIGHT UPPER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RIGHT LOWER
NNNNNNNNNNNNNNNNNNNNNNNNNN
The selected quadrant will appear in the brackets under
SEL QUAD .

Figure 4-7. Plot Quadrants
PLOT .
SEL QUAD and choose another quadrant where you want to place the displayed
measurement.
measurements .
Printing, Plotting, and Saving Measurement Results 4-17

If Y ou are Plotting to an HPGL Compatible Printer
Plot Function" located earlier in this chapter .
PLOT
NNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PLOTTER FORM FEED to print the data the printer has received.
Use test sequencing to automatically plot all four S-parameters .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQ
NEW SEQ/MODIFY SEQ
SEQUENCE 1 SEQ1
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MEAS
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNN
COPY
SEL QUAD
LEFT UPPER
PLOT
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MEAS
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNN
COPY
SEL QUAD
LEFT LOWER
PLOT
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNN
SEL QUAD
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RIGHT UPPER
NNNNNNNNNNNNNN
PLOT
COPY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
COPY
NNNNNNNNNNNNNNNNNNNNNNNNNN
SEL QUAD
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RIGHT LOWER
NNNNNNNNNNNNNN
PLOT
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQ
DONE SEQ MODIFY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQ
DO SEQUENCE
SEQUENCE 1 SEQ1
4-18 Printing, Plotting, and Saving Measurement Results

Plotting a Measurement to Disk
The plot les that you generate from the analyzer , contain the HPGL representation of the
measurement display . The les will not contain any setup or formfeed commands .
a.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNN
DISK .
SET ADDRESSES
PLOTTER PORT
b.
SELECT DISK and select the disk drive that you will plot to .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
INTERNAL DISK if you will plot to the analyzer internal disk drive .
EXTERNAL DISK if you will plot to a disk drive that is external to the analyzer .
Then con gure the disk drive as follows:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CONFIGURE EXT DISK
disk drive (default is 00) followed by
.
DISK UNIT NUMBER and enter the drive where your disk is located,
LOCAL
.
If your storage disk is partitioned, press
VOLUME NUMBER and enter the volume
number where you want to store the instrument state le .
PLOT .
The analyzer assigns the rst available default lename for the displayed directory . F or
example , the analyzer would assign PLOT00FP for a LIF format (PLOT00.FP for a DOS
format) if there were no previous plot les saved to the disk.
The gure below shows the three parts of the le name that is generated automatically by
the analyzer whenever a plot is requested. The two digit sequence number is incremented
by one each time a le with a default name is added to the directory .

Figure 4-8. A utomatic File Naming Convention for LIF F ormat
Printing, Plotting, and Saving Measurement Results 4-19

T o Output the Plot Files
Y ou can plot the les to a plotter from a personal computer .
Y ou can output your plot les to an HPGL compatible printer , by following the sequence
in \Outputting Plot Files from a PC to an HPGL Compatible Printer" located later in this
chapter .
Y ou can run a program that plots all of the les , with the root lename of PLOT , to an
HPGL compatible printer . This program is provided on the \Example Programs Disk" that is
included in the HP 8753D Network Analyzer Programmer's Guide. However , this program is
for use with LIF formatted disks and is written in HP B ASIC.
T o View Plot Files on a PC
Plot les can be viewed and manipulated on a PC using a word processor or graphics
presentation program. Plot les contain a text stream of HPGL (Hewlett-P ackard Graphics
Language) commands . In order to import a plot le into an application, that application must
have an import lter for HPGL (often times call HGL). Two such applications from the Lotus
suite of products are the word processor \AmiPro" and the graphics presentation package
\Freelance Graphics ."
Lotus applications are not supported by Hewlett-P ackard. The following
procedures are provided for informational use only . Other applications or other
versions of the same application may function di erently .
When viewed in such programs , the color and font size of the plot may vary from the output
of an HPGL/2 compatible color printer . The following table shows the di erences between
the color assignments of HPGL/2 compatible printers and Lotus applications . Also refer to
\Selecting P en Numbers and Colors" located earlier in this chapter .
HPGL/2 Printer Lotus Applications
No .
No .
3 blue 3 green
4 yellow 4 yellow
5 green 5 blue
6 red 6 red-violet
7 black 7 aqua
T o modify the color or font size , consult the documentation for the particular application being
used.
4-20 Printing, Plotting, and Saving Measurement Results

T o view plot les in AmiPro , perform the following steps:
le sux in the lename box to *.PL T .
The network analyzer does not use the sux *.PL T , so you may want to change
the lename lter to *.* or some other pattern that will allow you to locate the
les you wish to import.
colors . Y ou will probably need to change pen colors .
The network analyzer uses pen 7 for text. The default color in Ami Pro for pen
7 is aqua, which is not very readable against the typical white background. Y ou
may want to change pen 7 to black.
frame which can be sized to the page by grabbing one of the nodes and stretching the box as
required.
Y ou will notice that the annotation around the display is not optimum, as the Ami Pro
lter does not accurately import the HPGL command to render text.
Printing, Plotting, and Saving Measurement Results 4-21
Using AmiPro

T o view plot les in Freelance , perform the following steps:
The network analyzer does not use the sux *.HGL, so you may want to
change the lename lter to *.* or some other pattern that will allow you to
locate the les you wish to import.
Y ou will notice that when the trace is displayed, the text annotation will be illegible . Y ou
can easily x this with the following steps:
a. From the TEXT pull-down menu select FONT .
b. Select the type face and size . (F ourteen point text is a good place to start.)
c. Click OK to resize the font.
If you wish to modify the color of the displayed text, perform the following steps:
a. From the ARRANGE pull-down menu select UNGROUP .
b. Highlight a piece of text.
c. From the STYLE pull-down menu select A TTRIBUTES.
d. Select the desired text color and click OK.
e. Repeat steps b through d for each piece of text.
Outputting Plot Files from a PC to a Plotter
4-22 Printing, Plotting, and Saving Measurement Results
Using Freelance

Outputting Plot Files from a PC to an HPGL Compatible Printer
T o output the plot les to an HPGL compatible printer , you can use the HPGL initialization
sequence linked in a series as follows:
Do not insert spaces or linefeeds . Most editors allow the inclusion of escape sequences .
F or example , in the MS-DOS editor (DOS 5.0 or greater), press CNTRL-P (hold down the
CTRL key and press P) followed by the ESCape key to create the escape character .
T able 4-6. HPGL Initialization Commands
E conditional page eject
&12A page size 8.5 x 11
&l1O landscape orientation
(lower case l, one , capital O)
&a0L no left margin
(a, zero , capitol L)
&a400M no right margin
(a, 4, zero , zero , capitol M)
&l0E no top margin
(lower case l, zero , capitol E)
*c7680x5650Y frame size 10.66\x 7.847"
*p50x50Y move cursor to anchor point
*c0T set picture frame anchor point
*r-3U set CMY palette
%1B enter HPGL mode; cursor at PCL
As shown in T able 4-6 , the is the symbol used for the escape character ,
decimal value 27.
Printing, Plotting, and Saving Measurement Results 4-23
Command Remark · (720 decipoints/inch)

Do not insert spaces or linefeeds .
T able 4-7. HPGL T est File Commands
%0A exit HPGL mode
E form feed
Outputting Single P age Plots Using a Printer
Y ou can output plot les to an HPGL compatible printer , using the DOS command line and the
les created in the previous steps . This example assumes that the escape sequence les and the
plot les are in the current directory and the selected printer port is PRN.
C:> type hpglinit > PRN
C:> type PLOT00.FP > PRN
C:> type exithpgl > PRN
4-24 Printing, Plotting, and Saving Measurement Results
Command Remark · Command Remark

Outputting Multiple Plots to a Single P age Using a Printer
Refer to the \Plotting Multiple Measurements P er P age Using a Disk Drive ," located earlier in
this chapter , for the naming conventions for plot les that you want printed on the same page .
Y ou can use the following batch le to automate the plot le printing. This batch le must be
saved as \do plot.bat."
rem Description:
rem output HPGL initialization sequence to a le:spooler
rem append all the requested plot les to the spooler
rem append the formfeed sequence to the spooler
rem copy the le to the printer
rem (This routine uses COPY instead of PRINT because COPY
rem will not return until the action is completed. PRINT
rem will queue the le so the subsequent DEL will likely
rem generate an error . COPY avoids this .)
type hpglinit > spooler
for %%i in (%1) do type %%i >> spooler
type exithpgl >> spooler
F or example , you have the following list of les to plot:
PLOT00.LU
PLOT00.RL
PLOT00.RU
Y ou would invoke the batch print as follows:
C:> do_plot PLOT00.*
Printing, Plotting, and Saving Measurement Results 4-25

Plotting Multiple Measurements P er P age From Disk
The following procedures show you how to store plot les on a LIF formatted disk. A naming
convention is used so you can later run an HP B ASIC program on an external controller that
will output the les to the following peripherals:
a plotter with auto-feed capability , such as the HP 7550B
an HP-GL/2 compatible printer , such as the LaserJet 4 series (monochrome) or the DeskJet
1200C or DeskJet 1600C (color)
The program is contained on the \Example Programs Disk" that is provided with the
HP 8753D Network Analyzer Programmer's Guide. The le naming convention allows the
program to initiate the following:
to initialize the printer for HP-GL/2 at the beginning of a page
to plot multiple plot les on the same page
to send a page eject (form feed) to the hardcopy device , when all plots to the same page have
The plot le name is made up of four parts , the rst three are generated automatically by the
analyzer whenever a plot is requested. The two digit sequence number is incremented by one
each time a le with a default name is added to the directory .

Figure 4-9. Plot Filename Convention
T o Plot Multiple Measurements on a Full P age
Y ou may want to plot various les to the same page , for example , to show measurement data
traces for di erent input settings , or parameters , on the same graticule .
COPY
PLOT . The analyzer assigns the rst available default lename for the
displayed directory . F or example , the analyzer would assign PLOT00FP if there were no
previous plot les on the disk.
and turn the front panel knob to high-light the name of the le that
you just saved.
FILE UTILITIES
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RENAME FILE and turn the front panel knob to place the " pointer
to the A character .
4-26 Printing, Plotting, and Saving Measurement Results

SELECT LETTER
NNNNNNNNNNNNNN
DONE .
F or example , you may want only the data trace to appear on the second plot for
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
measurement comparison. In this case , you would press
DEFINE PLOT and choose
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PLOT DATA ON
PLOT MEM OFF
PLOT GRAT OFF
PLOT TEXT OFF
PLOT MKR OFF .
COPY
PLOT . The analyzer will assign PLOT00FP because you renamed the last le
saved.
and turn the front panel knob to high-light the name of the le that
you just saved.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FILE UTILITIES
RENAME FILE and turn the front panel knob to place the " pointer
to the B character .
SELECT LETTER
NNNNNNNNNNNNNN
DONE .
that you want to put on the same page . Renaming the les as shown below allows you to
use the provided program, that organizes and plots the les , according to the le naming
convention.
Plot File Recognized Filename
First File Saved PLOT00FP A
Second File Saved PLOT00FPB
Third File Saved PLOT00FPC
F ourth File Saved PLOT00FPD
The gure below shows plots for both the frequency and time domain responses of the same
device .

Figure 4-10. Plotting Two Files on the Same P age
Printing, Plotting, and Saving Measurement Results 4-27

T o Plot Measurements in P age Quadrants
SEL QUAD .
hardcopy . The selected quadrant appears in the brackets under
SEL QUAD .
Figure 4-11. Plot Quadrants
PLOT . The analyzer assigns the rst available default lename for the selected
quadrant. F or example , the analyzer would assign PLOT01LU if there were no other left
upper quadrant plots on the disk.
page . If you want to see what quadrants you have already saved, press
view the directory .
4-28 Printing, Plotting, and Saving Measurement Results

Titling the Displayed Measurement
Y ou can create a title that is printed or plotted with your measurement result.
NNNNNNNNNNNNNNNNN
DISPLA Y
MORE
TITLE to access the title menu.
ERASE TITLE .
SELECT LETTER .
a title that has a maximum of 50 characters . Press
BACK SPACE if you enter an incorrect
character .
DONE to complete the title entry .
Titles may also be entered using the optional external keyboard.
NNNNNNNNNNNNNNNNNNNNNNNNNN
FORMFEED keys are not intended for creating display titles .
Those keys are for creating commands to send to peripherals during a sequence
program.
Printing, Plotting, and Saving Measurement Results 4-29

Con guring the Analyzer to Produce a Time Stamp
Y ou can set a clock, and then activate it, if you want the time and date to appear on your
hardcopies .
SYSTEM
SET CLOCK .
SET YEAR and enter the current year (four digits), followed by
.
SET MONTH and enter the current month of the year , followed
.
SET DAY and enter the current day of the month, followed by
.
SET HOUR and enter the current hour of the day (0-23), followed by
.
SET MINUTES and enter the next immediate minute , followed by
.
ROUND SECONDS when the current time is exactly as you have set it.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TIME STAMP ON appears on the softkey label.
Aborting a Print or Plot Process
key to stop all data transfer .
again or reset the peripheral.
Printing or Plotting the List V alues or Operating P arameters
NNNNNNNNNNNNNN
LIST and select the information that you want to appear on your hardcopy .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LIST VALUES if you want a tabular listing of the measured data points , and
their current values , to appear on your hardcopy . This list will also include the limit test
information, if you have the limits function activated.
OP PARMS (MKRS etc) if you want a tabular listing of the parameters for both
measurement channels to appear on your hardcopy . The parameters include: operating
parameters , marker parameters , and system parameters that relate to the control of
peripheral devices .
If Y ou want a Single P age of V alues
PRINT MONOCHROME for a printer or
PLOT for a plotter peripheral, to create a
hardcopy of the displayed page of listed values .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NEXT PAGE to display the next page of listed values . Press
PREVIOUS PAGE to display
the previous page of listed values . Or , you can press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PREVIOUS PAGE
repeatedly to display a particular page of listed values that you want to appear on your
hardcopy . Then repeat the previous step to create the hardcopy .
4-30 Printing, Plotting, and Saving Measurement Results

listed values .
If you are printing the list of measurement data points , each page contains 30 lines of data.
The number of pages is determined by the number of measurement points that you have
selected under the
key .
If Y ou W ant the Entire List of V alues
PRINT ALL to print all pages of the listed values .
If you are printing the list of operating parameters , only the rst four pages are
printed. The fth page , system parameters , is printed by displaying that page
PRINT .
Printing, Plotting, and Saving Measurement Results 4-31
Choose · and then pressing

Solving Problems with Printing or Plotting
If you encounter a problem when you are printing or plotting, check the following list for
possible causes:
Look in the analyzer display message area. The analyzer may show a message that will
identify the problem. Refer to the \Error Messages" chapter if a message appears .
If necessary , refer to the con guration procedures in this chapter to check that you have
done the following:
connected an interface cable between the peripheral and the analyzer
connected the peripheral to ac power
switched on the power
switched the peripheral on line
selected the correct printer or plotter type
If you are using a laser printer for plotting, and the printer is outputting partial plots , the
printer may require more memory and/or the page protection activated.
Consult your printer manual for information on upgrading memory and how to
activate page protection.
Make sure that the analyzer address setting for the peripheral corresponds to the actual
HP-IB address of the peripheral. The procedure is explained earlier in this chapter .
Make sure that the analyzer is in system controller mode , by pressing
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SYSTEM CONTROLLER , if the analyzer is not connected to an external controller . Otherwise ,
the analyzer must be in the pass control mode .
Substitute the interface cable .
Substitute a di erent printer or plotter .
4-32 Printing, Plotting, and Saving Measurement Results

Saving and Recalling Instrument States
Places Where Y ou Can Save
oppy disk using the analyzer's internal disk drive
oppy disk using an external disk drive
IBM compatible personal computer using HP-IB mnemonics
What Y ou Can Save to the Analyzer's Internal Memory
The number of registers that the analyzer allows you to save depends on the size of associated
error-correction sets , and memory traces . Refer to the \Preset State and Memory Allocation"
chapter for further information.
Y ou can save instrument states in the analyzer internal memory , along with the following list of
analyzer settings . The default lenames are REG.
error-corrections on channels 1 and 2
When the ac line power is switched o , the internal non-volatile memory is
retained by a battery . The data retention time with the 3 V , 1.2 Ah battery is
as follows:
T emperature at 70
C : 250 days (0.68 year) characteristically
T emperature at 40
C : 1244 days (3.4 years) characteristically
T emperature at 25
C : 10 years characteristically
What Y ou Can Save to a Floppy Disk
Y ou can save an instrument state and/or measurement results to a disk. The default lenames
are FILEn, where n gets incremented by one each time a le with a default name is added to
the directory . The default lenames for data-only les are D A T AnDn (D A T An.Dn for DOS),
where the rst n is incremented by one each time a le with a default name is added to the
directory . The second n is the channel where the measurement was made . When you save a
le to disk, you can choose to save some or all of the following:
all settings listed above for internal memory
active error-correction for the active channel only
displayed measurement data trace
Printing, Plotting, and Saving Measurement Results 4-33

What Y ou Can Save to a Computer
Instrument states can be saved to and recalled from an external computer (system controller)
using HP-IB mnemonics . F or more information about the speci c analyzer settings that can be
saved, refer to the output commands located in the \Command Reference" chapter of the
HP 8753D Network Analyzer Programmer's Guide. F or an example program, refer to \Saving
and Recalling Instruments States" in the \Programming Examples" chapter of the
HP 8753D Network Analyzer Programmer's Guide.
4-34 Printing, Plotting, and Saving Measurement Results

Saving an Instrument State
SA VE/RECALL
SELECT DISK and select one of the storage devices:
INTERNAL MEMORY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
INTERNAL DISK
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
EXTERNAL DISK and then con gure as follows:
a. Connect an external disk drive to the analyzer's HP-IB connector , and con gure as
follows:
b. Press
DISK UNIT NUMBER and enter the drive where your disk is located,
.
If your storage disk is partitioned, press
VOLUME NUMBER and enter the volume number
where you want to store the instrument state le .
d.
SET ADDRESSES
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ADDRESS: DISK .
e. Enter the HP-IB address of the peripheral, if the default address is incorrect
(default = 00). F ollow the entry by pressing
.
f. Press
and select one of the following:
SYSTEM CONTROLLER to allow the analyzer to control peripherals directly .
TALKER/LISTENER to allow the computer controller to be involved in all
peripheral access operations .
USE PASS CONTROL to allow yourself to control the analyzer over HP-IB and
also allows the analyzer to take or pass control.
SA VE/RECALL
SAVE STATE .
The analyzer saves the state in the next available register , if you are saving to internal
memory , or saves the state to disk. Although one le is shown to represent an instrument
state on the analyzer display , each instrument state is composed of numerous les (which
can be viewed on a PC).
If you have saved enough les that you have used all the default names
(FILE00 - FILE31 for disk les , or REG1 - REG31 for memory les), you must
do one of the following in order to save more states:
rename an existing le to make a default name available
delete an existing le/register
Printing, Plotting, and Saving Measurement Results 4-35

Instrument states combined with measurements results can only be saved to disk. Files that
contain data-only , and the various save options available under the
DEFINE DISK SAVE key ,
are also only valid for disk saves .
The analyzer stores data in arrays along the processing ow of numerical data, from IF
detection to display . These arrays are points in the ow path where data is accessible , usually
via HP-IB . Y ou can choose from three di erent arrays which vary in modi cation exibility
when they are recalled.
data (raw data with error-correction applied)
format (data processed to the display format)
If you choose to save the raw data array , you will have the most exibility in modifying the
recalled measurement (including the ability to view all four S-parameters). This is because the
raw data array has the least amount of processing associated with it. Conversely , if you choose
to save the format array , your modi cation of the recalled measurement will be limited by all
the processes that are associated with that measurement result. However , the format array is
appropriate if you want to retrieve data traces that look like the currently displayed data.
De ne Save Modi cation Flexibility
Raw Data Array Most
F ormat Array Least
Y ou can also save data-only . This is saved to disk with default lenames D A T A00D1 to
D A T A31D1, for channel 1, or D A T A00D2 to D A T A31D2, for channel 2. However , these les are
not instrument states and cannot be recalled.
4-36 Printing, Plotting, and Saving Measurement Results


Figure 4-12. Data Processing Flow Diagram
If the analyzer has an active two-port measurement calibration, all
four S-parameters will be saved with the measurement results . All four
S-parameters may be viewed if the raw data array has been saved.
Measurement," located earlier in this chapter .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SA VE/RECALL
SELECT DISK .
INTERNAL DISK
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
EXTERNAL DISK and then con gure as follows:
a. Connect an external disk drive to the analyzer's HP-IB connector , and con gure as
follows:
b. Press
LOCAL
DISK UNIT NUMBER and enter the drive where your disk is located,
.
If your storage disk is partitioned, press
VOLUME NUMBER and enter the volume number
where you want to store the instrument state le .
d.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SET ADDRESSES
ADDRESS: DISK .
e. Enter the HP-IB address of the peripheral, if the default address is incorrect
(default = 00). F ollow the entry by pressing
.
Printing, Plotting, and Saving Measurement Results 4-37

f. Press
and select one of the following:
SYSTEM CONTROLLER to allow the analyzer to control peripherals directly .
TALKER/LISTENER to allow the computer controller to be involved in all
peripheral access operations .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PASS CONTROL to allow yourself to control the analyzer over HP-IB and also
allows the analyzer to take or pass control.
SA VE/RECALL
DEFINE DISK-SAVE .
DATA ARRAY ON
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RAW ARRAY ON
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FORMAT ARRAY ON
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
GRAPHICS ON
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DATA ONLY ON (see note below)
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DATA ARRAY ON ,
RAW ARRAY ON , or
FORMAT ARRAY ON , the data is stored
to disk in IEEE-64 bit real format (for LIF disks), and 32 bit PC format for DOS disks . This
makes the DOS data les half the size of the LIF les .
DATA ARRAY ON may store data to disk in the S2P ASCII data format.
See \ASCII Data F ormats ."
GRAPHICS ON , the user graphics area is saved. (Refer to the
HP 8753D Network Analyzer Programmer's Guide for information on using display
graphics .) The measurement display is not saved with this selection. (Refer to the
information located earlier in this chapter for a procedure that shows you how to plot
measurement displays to disk.)
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DATA ONLY ON will override all of the other save options . Because
this type of data is only intended for computer manipulation, the le contents
DATA ONLY ON save cannot be recalled and displayed on the analyzer .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SAVE USING BINARY for all applications except CITI le , S2P , or CAE applications .
SAVE USING ASCII for CITI le , S2P , and CAE applications or when you want to
import the information into a spread sheet format.
NNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RETURN
SAVE STATE .
4-38 Printing, Plotting, and Saving Measurement Results

ASCII Data F ormats
CITI le (Common Instrumentation Transfer and Interchange le) is an ASCII data format
that is useful when exchanging data between di erent computers and instruments . F or more
information on the CITI le data format as well as a list of CITI le keywords , refer to Appendix
A, \The CITI le Data F ormat and Keyword Reference ."
S2P Data F ormat
Component data les contain small signal S-parameters described by frequency dependent
linear network parameters for 2 port components . These les are assigned a lename with the
sux S1 or S2 depending on which measurement channel generated the data. These les are
output only . (They cannot be read in by the analyzer .)
An S2P le is output only when the user has either a full 2-port or TRL 2-port error correction
turned on, and has selected it under
:
DEFINE DISK-SAVE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DATA ARRAY ON or
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SAVE USING ASCII
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SAVE STATE
The template for component data les is as follows:
#
! indicates that all following on this line is a comment
# indicates that entries following on this line are parameters that are being
GHz, MHz, kHz, Hz
parameter S for S-parameters
format DB for dB magnitude and angle in degrees
MA for linear magnitude and angle in degrees
RI for real and imaginary pair
Rn the reference impedance in ohms for the analyzer making the measurement
(R 50 or R 75)
The \format" above is selected by the current selection under the FORMA T menu. T o select the
DB format, the FORMA T must be LOG MA G. F or MA, the FORMA T must be LIN MA G, and all
other FORMA T selections will output RI data. The S2P data will always represent the format
array data, including e ects of electrical delay and port extensions . A CITI le will be saved at
the same time . T o be consistent with previous versions , the CITI le data saved will represent
the D A T A array (corrected data) without e ects of electrical delay or port extensions .
Printing, Plotting, and Saving Measurement Results 4-39

Here is an S2P example le for an 21 point measurement of a 20 dB attenuator:
! Network Analyzer HP 8753D.06.11 Serial No. US31240052
!
! 23 May 1997 15:26:54
50000000 56.404 145.38 .0083 .3337 .0079 .1606 58.034 5.0084
1050000000 68.761 65.356 .0142 .0137 .0042 .1043 64.085 83.573
2050000000 64.108 41.723 .0253 .0068 .0147 .1675 61.954 173.75
3050000000 60.125 119.38 .0358 0.0 .0279 .1455 60.338 56.346
4050000000 61.224 32.686 .0474 .0137 .0384 .1249 61.743 169.73
5050000000 59.429 38.486 .0596 .0494 .0448 .0700 55.876 156.44
6050000000 56.035 70.648 .0681 .0975 .0553 .0315 63.449 39.47
7050000000 54.229 88.746 .0749 .1139 .0633 .0068 55.804 30.247
8050000000 61.411 111.97 .0802 .1977 .0712 .0521 51.102 97.546
9050000000 52.49 103.21 .0828 .2952 .0764 .1249 52.406 126.36
10050000000 64.291 35.461 .0875 .3213 .0775 .2252 59.417 85.038
11050000000 52.096 46.505 .0918 .4298 .0770 .2774 48.868 68.46
12050000000 49.648 78.573 .0878 .5232 .0787 .3364 50.699 77.157
13050000000 48.431 25.793 .0805 .5616 .0751 .4229 48.461 60.445
14050000000 45.984 36.612 .0717 .6097 .0651 .4202 44.971 37.711
15050000000 52.703 9.3823 .0748 .6001 .0614 .3749 46.822 23.128
16050000000 50.548 63.182 .0863 .5685 .0849 .3364 53.049 11.283
17050000000 57.776 19.931 .0973 .5877 .0971 .4229 48.105 34.254
18050000000 56.256 98.687 .1022 .7045 .0993 .5081 54.446 67.992
19050000000 76.33 149.78 .0965 .7635 .1004 .5644 48.489 48.591
20050000000 59.269 163.78 .1050 .7951 .1078 .6083 44.865 2.8304
4-40 Printing, Plotting, and Saving Measurement Results

Re-Saving an Instrument State
If you re-save a le , the analyzer overwrites the existing le contents .
Y ou cannot re-save a le that contains data only . Y ou must create a new le .
SELECT DISK and select the storage device .
INTERNAL MEMORY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
INTERNAL DISK
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
EXTERNAL DISK
NNNNNNNNNNNNNNNNNNNN
RETURN and then use the
* keys or the front panel knob to high-light the name
of the le that you want to re-save .
NNNNNNNNNNN
RE-SAVE STATE
YES .
SA VE/RECALL
SELECT DISK .
INTERNAL MEMORY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
INTERNAL DISK
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
EXTERNAL DISK
NNNNNNNNNNNNNNNNNNNN
RETURN .
T o Delete an Instrument State File
* keys or the front panel knob to high-light the name of the le that you want
to delete .
FILE UTILITIES
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DELETE FILE
NNNNNNNNNNN
YES to delete all of the les that make up the
selected instrument state .
T o Delete all Files
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNN
FILE UTILITIES
DELETE ALL FILES
YES to delete all of the les that are on the
selected storage device .
Printing, Plotting, and Saving Measurement Results 4-41

SA VE/RECALL
SELECT DISK .
INTERNAL MEMORY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
INTERNAL DISK
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
EXTERNAL DISK
N
RETURN and then use the
* keys or the front panel knob to high-light the name
of the le that you want to rename .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RETURN
FILE UTILITIES
RENAME FILE
ERASE TITLE .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SELECT LETTER when the arrow points to each character . Press
BACK SPACE if you enter
an incorrect character . After you have selected all the characters in the new lename , press
DONE .
Renaming les may also be done by using the optional external keyboard.
SA VE/RECALL
SELECT DISK .
INTERNAL MEMORY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
INTERNAL DISK
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
EXTERNAL DISK
* keys or the front panel knob to high-light the name of the le that you
want to recall.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RECALL STATE .
4-42 Printing, Plotting, and Saving Measurement Results

F ormatting a Disk
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SA VE/RECALL
FILE UTILITIES
FORMAT DISK .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FORMAT:LIF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FORMAT:DOS
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNN
FORMAT EXT DISK
YES .
Solving Problems with Saving or Recalling Files
If you encounter a problem when you are storing les to disk, or the analyzer internal memory ,
check the following list for possible causes:
Look in the analyzer display message area. The analyzer may show a message that will
identify the problem. Refer to the \Error Messages" chapter if you view a message .
Make sure that you are NOT using a single-sided oppy disk in the analyzer disk drive .
Make sure that you are using a formatted disk.
Make sure that the disk has not been formatted with the LIF-1 (hierarchical le system)
extensions as the analyzer does not support this format.
If Y ou are Using an External Disk Drive
Make sure that the analyzer is in system controller mode , by pressing
N
SYSTEM CONTROLLER .
Make sure that you have connected the disk drive to ac power , switched on the power , and
connected an HP-IB cable between the disk drive and the analyzer .
Make sure that the analyzer recognizes the disk drive's HP-IB address , as explained earlier in
this chapter .
Make sure that the analyzer recognizes the disk (drive) unit that you selected (0 or 1).
If the external disk is a hard disk, make sure that the disk volume number is set correctly .
If the disk drive is an older HP 9122, it may not recognize the newer high density disks .
Substitute the HP-IB cable .
Substitute the disk drive .
Printing, Plotting, and Saving Measurement Results 4-43



This chapter describes techniques and analyzer functions that help you achieve the best
measurement results . The following topics are included in this chapter:
P erformance veri cation
Reference plane and port extensions
Frequency response and isolation correction
One-port re ection correction
TRL* and TRM* error-correction
Modifying calibration kit standards
P ower meter measurement calibration
Calibrating for noninsertable devices
Modify the cal kit thru de nition
Maintaining testport output power during sweep retrace
Making accurate measurements of electrically long devices
Optimizing Measurement Results 5-1

Where to Look for More Information
A dditional information about many of the topics discussed in this chapter is located in the
following areas:
Chapter 2, \Making Measurements ," contains step-by-step procedures for making
measurements or using particular functions .
Chapter 4, \Printing, Plotting, and Saving Measurement Results ," contains instructions
for saving to disk or to the analyzer internal memory , and printing and plotting displayed
measurements .
Chapter 6, \Application and Operation Concepts ," contains explanatory-style information
about many applications and analyzer operation.
Increasing Measurement A ccuracy
Connector repeatability is a source of random measurement error . Measurement
error-corrections do not compensate for these errors . F or all connectors , you should frequently
do the following:
use correct connection techniques (See Chapter 2, T able 2-1.)
Cables connecting the device under test to the analyzer can contribute random errors to your
measurement. Y ou should frequently do the following:
inspect for lossy cables
inspect for damaged cable connectors
practice good connector care techniques
minimize cable position changes between error-correction and measurements
inspect for cables which dramatically change magnitude or phase response when exing (This
may indicate an intermittent problem.)
Electrical characteristics will change with temperature due to the thermal expansion
characteristics of devices within the analyzer , calibration devices , test devices , cables , and
adapters . Therefore , the operating temperature is a critical factor in their performance . During
a measurement calibration, the temperature of the calibration devices must be stable and
within 25 5
use a temperature-controlled environment
ensure the temperature stability of the calibration devices
avoid handling the calibration devices unnecessarily during calibration
ensure the ambient temperature is 1
C of measurement error-correction temperature
5-2 Optimizing Measurement Results

Minute changes in frequency accuracy and stability can occur as a result of temperature and
aging (on the order of parts per million). If you require greater frequency accuracy , do the
following:
Override the internal crystal with a high-stability external source , frequency standard, or (if
your analyzer is equipped with Option 1D5) use the internal frequency standard.
P erformance V eri cation
Y ou should periodically check the accuracy of the analyzer measurements , by doing the
following:
perform a measurement veri cation at least once per year
Refer to the HP 8753D Service Guide for the measurement veri cation procedure .
Reference Plane and P ort Extensions
Use the port extension feature to compensate for the phase shift of an extended measurement
reference plane , due to such additions as cables , adapters , and xtures , after completing an
error-correction procedure (or when there is no active correction).
Using port extensions is similar to using electrical delay . However , using port extensions is the
preferred method of compensating for test xture phase shift. T able 5-1 explains the di erence
between port extensions and electrical delay .
T able 5-1.
Di erences between PORT EXTENSIONS and ELECTRICAL DELA Y
PORT EXTENSIONS
F
FFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFF
ELECTRICAL DELAY
Main E ect The end of a cable becomes the test port plane
Compensates for the electrical length of a
for all S-parameter measurements .
cable .
Set the cable's electrical length x 1 for
transmission.
Set the cable's electrical length x 2 for
re ection.
All S-parameters . Only the currently selected S-parameter .
Intelligently compensates for 1 times or 2
Only compensates for electrical length.
times the cable's electrical delay , depending
on which S-parameter is computed.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Y ou can activate a port extension by pressing
MORE
PORT EXTENSIONS
EXTENSIONS ON .
Then enter the delay to the reference plane .
Optimizing Measurement Results 5-3

The accuracy of network analysis is greatly in uenced by factors external to the network
analyzer . Components of the measurement setup , such as interconnecting cables and adapters ,
introduce variations in magnitude and phase that can mask the actual response of the device
under test.
Error-correction is an accuracy enhancement procedure that removes systematic errors
(repeatable measurement variations) in the test setup . The analyzer measures known standard
devices , and uses the results of these measurements to characterize the system.
Conditions Where Error-Correction is Suggested
Measurement accuracy and system characteristics can be a ected by the following factors:
A dapting to a di erent connector type or impedance .
Connecting a cable between the test device and an analyzer test port.
Connecting any attenuator or other such device on the input or output of the test device .
If your test setup meets any of the conditions above , the following system characteristics may
be a ected:
amplitude at device input
Several types of error correction are available that remove from one to twelve systematic
errors . The full 2-port correction e ectively removes all twelve correctable systematic errors .
Some measurements do not require correction for all twelve errors . The following table
explains each correction and its uses .
5-4 Optimizing Measurement Results

T able 5-2. Purpose and Use of Di erent Error-Correction Procedures
Response Transmission or re ection
Frequency response . Thru for transmission, open
measurement when the highest
or short for re ection.
accuracy is not required.
Response & isolation Transmission of high insertion loss
Same as response plus
devices or re ection of high return
isolation in transmission or
isolation standard. (load)
loss devices . Not as accurate as
directivity in re ection.
1-port or 2-port correction.
1-port Re ection of any one-port device
Directivity , source match,
Short and open and load.
or well terminated two-port
frequency response .
device .
1-port Re ection of any one-port device
Directivity , source match,
Short and open and load.
or well terminated two-port
frequency response .
device .
Full 2-port Transmission or re ection of
Directivity , source match,
Short and open and load and
highest accuracy for two-port
thru. (2 loads for isolation)
devices .
forward and reverse .
/LRM
Transmission or re ection when
Thru, re ect, line , or line ,
highest accuracy is not required.
frequency response .
re ect, match, or thru,
re ect, match.
Frequency response calibration is not as accurate as other calibration methods .
Error-correction is only valid for a speci c stimulus state , which you must select before you
start a correction. If you change any of the following parameters , you will invalidate the
correction and the analyzer will switch the correction o (unless the interpolated error
correction feature is activated):
The error-correction quality may be degraded (Cor changes to C? ), if you change the following
stimulus state parameters:
If you activate averaging, the analyzer may increase the sweep time if more
time is needed to calculate the averages . If the sweep time changes , you will
see Cor change to C? . The number of averages does not a ect a sweep cycle
time . Therefore , if you use averaging for error-correction, leave it on for the
measurement and set the averaging factor to 1, for a faster sweep .
Optimizing Measurement Results 5-5

The quality of the error-correction is limited by two factors: (1) the di erence between the
model of the calibration standards and the actual electrical characteristics of those standards ,
and (2) the condition of the calibration standards . T o make the highest quality measurement
calibration, follow the suggestions below:
use the correct standard model
inspect the calibration standards
clean the calibration standards
gauge the calibration standards
use correct connection techniques
If you want to use calibration standards other than the default sets , you must change the
standard model. (Refer to \Modifying Calibration Kit Standards" located later in this chapter .)
After you enter the mathematical model for the new calibration standards , the analyzer can
then use the model that corresponds to the new standards .
Compensating for the Electrical Delay of Calibration Standards
Short and open calibration standards in the 3.5 mm, 2.4 mm, and 2.92 mm connector types
have a certain amount of electrical delay . The analyzer compensates for this delay by o setting
the calibration results by the total amount of electrical delay caused by the calibration standard
in both the forward and reverse direction. As a result, if these standards are measured after a
calibration, they will not appear to be \perfect" shorts or opens . This is an indication that your
analyzer is working properly and that it has successfully performed a calibration.
If you enter the opposite amount of electrical delay that was used by the
analyzer during calibration, then the short calibration standard will appear
to be \perfect." The open calibration standard has additional phase shift
caused by fringing capacitance . See \Calibration Considerations" in Chapter 6,
\Application and Operation Concepts ."
Clarifying Type-N Connector Sex
When you are performing error-correction for a system that has type-N test port connectors ,
the softkey menus label the sex of the test port connector - not the calibration standard
connector . F or example , the label
SHORT (F) refers to the short that will be connected to the
female test port.
When to Use Interpolated Error-Correction
Y ou may want to use interpolated error-correction when you choose a subset of a frequency
range that you already corrected, when you change the number of points , or when you change
to CW . This feature also allows you to change the parameters in a 2-port correction, such as IF
bandwidth, power , or sweep time . The analyzer calculates the systematic errors from the errors
of the original correction.
The quality of the interpolated error-correction depends on the amount of phase shift and
amplitude change of the error coecients between measurement points . If the phase shift
per ve measurement points , the interpolated error-correction can be a great
improvement over uncorrected measurement.
T o activate interpolated measurement correction, press
INTERPOL ON
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CORRECTION ON .
When interpolation is in use , the notation C? will appear on the analyzer display .
5-6 Optimizing Measurement Results

Procedures for Error-Correcting Y our Measurements
This section has example procedures or information on the following topics:
frequency response and isolation correction
one-port re ection correction
modifying calibration kit standards
power meter measurement calibration procedure
If you are making measurements on uncoupled measurement channels , you
must make a correction for each channel.
Optimizing Measurement Results 5-7
frequency response correction · full two-port correction · TRL*/LRM* correction

Y ou can remove the frequency response of the test setup for the following measurements .
combined re ection and transmission measurements
Response Error-Correction for Re ection Measurements
.
If you want to make a re ection measurement on PORT 1 (in the forward direction, S
leave the instrument default setting.
If you want to make a re ection measurement on PORT 2 (in the reverse direction S
press:
Set any other measurement parameters that you want for the device measurement:
power , sweep type , number of points , or IF bandwidth.
T o access the measurement error-correction menus , press:
If your calibration kit is di erent than the kit speci ed under the
CAL KIT [ ] softkey ,
press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL KIT
SELECT CAL KIT (select your type of kit)
RETURN
If your type of calibration kit is not listed in the displayed menu, refer to the \Modifying
Calibration Standards" procedure , located later in this chapter .
T o select a response correction, press:
NNNNNNNNNNNNNNNNNNNNNNNNNN
CAL
CALIBRATE MENU
RESPONSE
Connect the short or open calibration standard to the port you selected for the test port
(PORT 1 for S
or PORT 2 for S
).
Include any adapters or cables that you will have in the device measurement.
That is , connect the standard device to the particular connector where you will
connect your device under test.
5-8 Optimizing Measurement Results


Figure 5-1.
Standard Connections for a Response Error-Correction for Re ection Measurement
T o measure the standard when the displayed trace has settled, press:
NNNNNNNNNNNNNN
OPEN
If the calibration kit you selected has a choice between male and female calibration
standards , remember to select the sex that applies to the test port and not the standard.
The analyzer displays WAIT - MEASURING CAL STANDARD during the standard measurement.
The analyzer underlines the softkey that you selected after it nishes the measurement,
and computes the error coecients .
This calibration allows only one standard to be measured. If you press the
wrong key for a standard, start over with step 6. Do not use a thru standard
for a re ection response correction.
Y ou can save or store the measurement correction to use for later
measurements , that use the same measurement parameters . Refer to the
\Printing, Plotting, and Saving Measurement Results" chapter for procedures .
This completes the response correction for re ection measurements . Y ou can connect and
measure your device under test.
Optimizing Measurement Results 5-9
SHORT or

Response Error-Correction for Transmission Measurements
.
If you want to make a transmission measurement in the forward direction (S
), press:
If you want to make a transmission measurement in the reverse direction (S
), press:
MEAS
number of points , IF bandwidth.
NNNNNNNNNNNNNNNNNNNNNNNNNN
CAL
CALIBRATE MENU
RESPONSE
test.
Include any adapters or cables that you will have in the device measurement.
That is , connect the standard device where you will connect your device under
test.

Figure 5-2.
Standard Connections for Response Error-Correction for Transmission Measurements
THRU
The analyzer displays WAIT - MEASURING CAL STANDARD during the standard measurement.
The analyzer underlines the
THRU softkey after it measures the calibration standard, and
computes the error coecients .
5-10 Optimizing Measurement Results

Do not use an open or short standard for a transmission response correction.
Y ou can save or store the measurement correction to use for later
measurements . Refer to the \Printing, Plotting, and Saving Measurement
Results" chapter for procedures .
measure your device under test.
Receiver calibration provides a frequency response error-correction that also indicates absolute
power in dBm. This calibration is most useful when performed with a power meter calibration.
later in this chapter .
POWER
MENU
test.
Include any adapters or cables that you will have in the device measurement.
That is , connect the standard device where you will connect your device under
test.

Figure 5-3. Standard Connections for Receiver Calibration
INPUT PORTS
NNNNN
B
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TEST PORT 1
MEAS
This sets the source at PORT 1, and the measurement receiver to PORT 2, or B channel.
number of points , IF bandwidth.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL
RECEIVER CAL
TAKE RCVR CAL SWEEP
Optimizing Measurement Results 5-11

Y ou can save or store the measurement correction to use for later
measurements . Refer to the \Printing, Plotting, and Saving Measurement
Results" chapter for procedures .
measure your device under test.
The accuracy of the receiver calibration will be nearly the same as the test
port power accuracy; and the test port power accuracy can be signi cantly
improved by performing a power meter source calibration, as described later in
5-12 Optimizing Measurement Results
this chapter

Frequency Response and Isolation Error-Corrections
removes frequency response of the test setup
removes isolation in transmission measurements
removes directivity in re ection measurements
Y ou can make a response and isolation correction for the following measurements .
combined re ection and transmission measurements
Response and Isolation Error-Correction for Re ection Measurements
Although you can perform a response and isolation correction for re ection measurements ,
Hewlett-P ackard recommends that you perform an S
one-port error-correction; it is more
accurate and just as convenient.
.
If you want to make a re ection measurement on PORT 1 (in the forward direction, S
leave the instrument default setting.
If you want to make a re ection measurement on PORT 2 (in the reverse direction, S
press:
MEAS
sweep type , number of points , IF bandwidth.
If your calibration kit is di erent than the kit speci ed under the
CAL KIT [ ] softkey ,
press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNN
CAL KIT
SELECT CAL KIT (select your type of kit)
If your type of calibration kit is not listed in the displayed menu, refer to the \Modifying
Calibration Kit Standards" procedure , located later in this chapter .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNN
CALIBRATE MENU
RESPONSE & ISOL'N
RESPONSE
(PORT 1 for S
or PORT 2 for S
).
Include any adapters that you will have in the device measurement. That is ,
connect the standard device to the particular connector where you will connect
your device under test.
Optimizing Measurement Results 5-13


Figure 5-4.
Standard Connections for a Response and Isolation Error-Correction for Re ection
NNNNNNNNNNNNNN
OPEN
If the calibration kit you selected has a choice between male and female calibration
standards , remember to select the sex that applies to the test port and not the standard.
The analyzer displays WAIT - MEASURING CAL STANDARD during the standard measurement.
The analyzer underlines the softkey that you selected after it nishes the measurement,
and computes the error coecients .
ISOL'N STD
DONE RESP ISOL'N CAL
The analyzer displays the corrected S
) data. The analyzer also shows the notation
Cor to the left of the screen, indicating that the correction is switched on for this channel.
Y ou can save or store the error-correction to use for later measurements .
Refer to the \Printing, Plotting, and Saving Measurement Results" chapter for
procedures .
can connect and measure your device under test.
5-14 Optimizing Measurement Results
Measurements · SHORT or · (or S

Response and Isolation Error-Correction for Transmission Measurements
This procedure is intended for measurements that have a measurement range of greater than
90 dB .
.
If you want to make a transmission measurement in the forward direction (S
), press:
MEAS
If you want to make a transmission measurement in the reverse direction (S
), press:
MEAS
number of points , IF bandwidth.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
If your calibration kit is di erent than the kit speci ed under the
CAL KIT [ ] softkey ,
press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL KIT
SELECT CAL KIT (select your type of kit)
If your type of calibration kit is not listed in the displayed menu, refer to the \Modifying
Calibration Kit Standards" procedure , located later in this chapter .
CALIBRATE MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RESPONSE & ISOL'N
NNNNNNNNNNNNNNNNNNNNNNNNNN
RESPONSE
test.
Include any adapters that you will have in the device measurement. That is ,
connect the standard device to the particular connector where you will connect
your device under test.
THRU
The analyzer displays WAIT - MEASURING CAL STANDARD during the standard measurement.
The analyzer underlines the
THRU softkey after it measures the calibration standard, and
computes the error coecients .
the adapters that you would include for your device measurement.
Optimizing Measurement Results 5-15


Figure 5-5.
Standard Connections for a Response and Isolation Error-Correction for Transmission
If you will be measuring highly re ective devices , such as lters , use the test
device , connected to the reference plane and terminated with a load, for the
isolation standard.
a.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
A VG
AVERAGING ON
AVERAGING FACTOR and enter at least four times more
averages than desired during the device measurement.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
b.
MORE
ALTERNATE A and B to eliminate one crosstalk path.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL
RESUME CAL SEQUENCE
ISOL'N STD
averaging factor by at least four times or turn averaging o .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL
RESUME CAL SEQUENCE
DONE RESP ISOL'N CAL
The analyzer displays the corrected data trace . The analyzer also shows the notation Cor at
the left of the screen, indicating that the correction is switched on for this channel.
Y ou can save or store the measurement correction to use for later
measurements . Refer to the \Printing, Plotting, and Saving Measurement
Results" chapter for procedures .
can connect and measure your device under test.
5-16 Optimizing Measurement Results

One-P ort Re ection Error-Correction
removes directivity errors of the test setup
removes source match errors of the test setup
removes frequency response of the test setup
Y ou can perform a 1-port correction for either an S
measurement. The only
di erence between the two procedures is the measurement parameter that you select.
This is the recommended error-correction process for all re ection
measurements , when full two-port correction is not used.
.
If you want to make a re ection measurement on PORT 1 (in the forward direction, S
leave the instrument default setting.
If you want to make a re ection measurement on PORT 2 (in the reverse direction, S
press:
number of points , IF bandwidth.
If your calibration kit is di erent than the kit speci ed under the
CAL KIT [ ] softkey ,
press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SELECT CAL KIT (select your type of kit)
CAL KIT
RETURN
If your type of calibration kit is not listed in the displayed menu, refer to the \Modifying
Calibration Kit Standards" procedure , located later in this chapter .
CALIBRATE MENU and select the correction type
If you want to make a re ection measurement at PORT 1 press:
S11 1-PORT
If you want to make a re ection measurement at PORT 2, press:
S22 1-PORT
measurement).
Include any adapters that you will have in the device measurement. That is ,
connect the calibration standard to the particular connector where you will
connect your device under test.
Optimizing Measurement Results 5-17


Figure 5-6. Standard Connections for a One P ort Re ection Error-Correction
N
OPEN
If the calibration kit that you selected has a choice between male or female
calibration standards , remember to select the sex that applies to the test port
and not the standard.
The analyzer displays WAIT - MEASURING CAL STANDARD during the standard measurement.
The analyzer underlines the
OPEN softkey after it measures the calibration standard.
NNNNNNNNNNNNNNNN
SHORT
NNNNNNNNNNNNNNNNN
The analyzer measures the short circuit and underlines the
SHORT softkey .
When the displayed trace settles , press
LOAD .
The analyzer measures the load and underlines the
LOAD softkey .
DONE: 1-PORT CAL
The analyzer displays the corrected data trace . The analyzer also shows the notation Cor to
the left of the screen, indicating that the correction is switched on for this channel.
The open, short, and load could be measured in any order , and need not follow
the order in this example .
5-18 Optimizing Measurement Results

Y ou can save or store the error-correction to use for later measurements .
Refer to the \Printing, Plotting, and Saving Measurement Results" chapter for
procedures .
measure your device under test.
Optimizing Measurement Results 5-19

Full Two-P ort Error-Correction
removes directivity errors of the test setup in forward and reverse directions
removes source match errors of the test setup in forward and reverse directions
removes load match errors of the test setup in forward and reverse directions
removes isolation errors of the test setup in forward and reverse directions (optional)
removes frequency response of the test setup in forward and reverse directions
This is the most accurate error-correction procedure . Since the analyzer takes
both forward and reverse sweeps , this procedure takes more time than the
other correction procedures .
format, number of points , IF bandwidth.
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
If your calibration kit is di erent than the kit speci ed under the
CAL KIT [ ] softkey ,
press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNN
CAL KIT
SELECT CAL KIT (select your type of kit)
If your type of calibration kit is not listed in the displayed menu, refer to the \Modifying
Calibration Kit Standards" procedure , located later in this chapter .
CALIBRATE MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FULL 2-PORT
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
REFLECTION
Include any adapters that you will have in the device measurement. That is ,
connect the standard to the particular connector where you will connect your
device under test.

Figure 5-7. Standard Connections for Full Two port Error-Correction
5-20 Optimizing Measurement Results
·

FORWARD: OPEN
The analyzer displays WAIT - MEASURING CAL STANDARD during the standard measurement.
The analyzer underlines the
OPEN softkey after it measures the standard.
FORWARD: SHORT
The analyzer measures the short circuit and underlines the
SHORT softkey .
FORWARD: LOAD
NNNNNNNNNNNNNN
The analyzer measures the load and underlines the
LOAD softkey .
to PORT 2, and use the
REVERSE: OPEN ,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
REVERSE: LOAD softkeys .
Include any adapters that you would include in your device measurement.
STANDARDS DONE
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
T o start the transmission portion of the correction, press:
test as shown in Figure 5-7.
Include any adapters or cables that you will have in the device measurement.
That is , connect the standard device where you will connect your device under
test.
Optimizing Measurement Results 5-21

The thru in most calibration kits is de ned with zero length. The correction
will not work properly if a non-zero length thru is used, unless the calibration
kit is modi ed to change the de ned thru to the length used. This is important
for measurements of noninsertable devices (devices having ports that are both
male or both female). The modi ed calibration kit must be saved as the user
calibration kit, and the
USER KIT must be selected before the calibration is
started.
DO BOTH FWD+REV
The analyzer underlines the softkey label after it makes each measurement.
ISOLATION and select from the following two options:
If you will be measuring devices with a dynamic range less than 90 dB , press:
OMIT ISOLATION
If you will be measuring devices with a dynamic range greater than 90 dB , follow these
steps:
a. Connect impedance-matched loads to PORT 1 and PORT 2. Include the adapters that
you would include for your device measurement.
If you will be measuring highly re ective devices , such as lters , use the test
device , connected to the reference plane and terminated with a load, for the
isolation standard.
b. A ctivate at least four times more averages than desired during the device
measurement.
If loads can be connected to both port 1 and port 2 simultaneously , then the
following step can be performed using the
DO BOTH FWD + REV softkey .
c. Press
RESUME CAL SEQUENCE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ISOLATION
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FWD ISOL'N ISOL'N STD
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
REV ISOL'N ISOL'N STD
ISOLATION DONE .
d. Return the averaging to the original state of the measurement, and press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RESUME CAL SEQUENCE .
DONE 2-PORT CAL
The analyzer displays the corrected measurement trace . The analyzer also shows the
Y ou can save or store the measurement correction to use for later
measurements . Refer to the \Printing, Plotting, and Saving Measurement
Results" chapter for procedures .
device under test.
5-22 Optimizing Measurement Results
Press · ·

TRL* and TRM* Error-Correction
The HP 8753D analyzer has the capability of making calibrations using the TRL*/LRM* method.
Y ou must have a TRL calibration kit de ned and saved in the
USER KIT , as shown in
\Modifying Calibration Kit Standards ," located later in this section.
NNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL
CAL KIT
SELECT CAL KIT
USER KIT
RETURN
RETURN
CALIBRATE MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TRL*/LRM* 2-PORT .
test ports .
THRU TRLTHRU
S11 REFL: TRLSHORT
S22 REFL: TRLSHORT
PORT 2.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DO BOTH FWD+REV .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
The line data is measured and the
labels are underlined.
ISOLATION
Y ou could choose not to perform the isolation measurement by pressing
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
OMIT ISOLATION
DONE TRL/LRM .
Y ou should perform the isolation measurement when the highest dynamic range
is desired.
T o perform the best isolation measurements , you should reduce the system
bandwidth, and/or activate the averaging function.
A poorly measured isolation class can actually degrade the overall
measurement performance . If you are in doubt of the isolation measurement
quality , you should omit the isolation portion of this procedure .
If loads can be connected to both port 1 and port 2 simultaneously , then the
following measurement can be performed using the
DO BOTH FWD + REV
softkey .
Optimizing Measurement Results 5-23

REV ISOL'N ISOL'N STD
FWD ISOL'N ISOL'N STD
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ISOLATION DONE
steps . When the analyzer detects that you have made all the necessary measurements , the
message line will show PRESS 'DONE' IF FINISHED WITH CAL . Press
DONE TRL/LRM .
The message COMPUTING CAL COEFFICIENTS will appear , indicating that the analyzer is
performing the numerical calculations of error coecients .
Y ou can save or store the measurement correction to use for later
measurements . Refer to the \Printing, Plotting, and Saving Measurement
Results" chapter for procedures .
Y ou must have a TRM calibration kit de ned and saved in the
USER KIT , as shown in
\Modifying Calibration Kit Standards ," located later in this section.
This must be done before performing the following sequence .
CAL KIT
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SELECT CAL KIT
NNNNNNNNNNNNNNNNNNNNNNNNNN
USER KIT
NNNNNNNNNNNNNNNNNNNN
RETURN
NNNNNNNNNNNNNNNNNNNN
RETURN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CALIBRATE MENU
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TRL*/LRM* 2-PORT .
test ports .
THRU TRMTHRU
S11 REFL: TRMSHORT
S22 REFL: TRMSHORT
If loads can be connected to both port 1 and port 2 simultaneously ,
then the following TRM load measurement can be performed using the
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DO BOTH FWD + REV softkey .
Refer to \Choosing Calibration Load Standards ."
5-24 Optimizing Measurement Results
TRM Error-Correction ·

LINE/MATCH
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LN/MATCH1 TRMLOAD to access the Loads menu. When the displayed
trace settles , press the softkey corresponding to the load used. If a sliding load is
SLIDING to access the Sliding Load menu. P osition the slide and press
SLIDE IS SET .
LN/MATCH1 TRMLOAD softkey label is underlined.
Connect the load to PORT 2 and press
LN/MATCH2 TRMLOAD .
DONE LINE/MATCH
ISOLATION
Y ou could choose not to perform the isolation measurement by pressing
OMIT ISOLATION
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DONE TRL/LRM .
Y ou should perform the isolation measurement when the highest dynamic range
is desired.
T o perform the best isolation measurements , you should reduce the system
bandwidth, and/or activate the averaging function.
A poorly measured isolation class can actually degrade the overall
measurement performance . If you are in doubt of the isolation measurement
quality , you should omit the isolation portion of this procedure .
steps . When the analyzer detects that you have made all the necessary measurements , the
message line will show PRESS 'DONE' IF FINISHED WITH CAL. Press
DONE TRL/LRM .
The message COMPUTING CAL COEFFICIENTS will appear , indicating that the analyzer is
performing the numerical calculations of error coecients .
Y ou can save or store the measurement correction to use for later
measurements . Refer to the \Printing, Plotting, and Saving Measurement
Results" chapter for procedures .
When making measurements using the same port with uncoupled channels , the
power level for each channel must fall within the same power range setting of
that single port. An error message will be displayed if you enter two power
levels that do not fall within the same power range .
Optimizing Measurement Results 5-25
Press · used, press · the

Modifying Calibration Kit Standards
Numerical data for most Hewlett-P ackard calibration kits is provided in the
calibration kit manuals .
The following section provides a summary of the information in the 8510-5A application note ,
as well as HP 8753D menu-speci c information. F or a detailed description of the menus and
softkeys located in this section, as well as information about when user-de ned calibration kits
should be used, refer to Chapter 6, \Application and Operation Concepts ."
The following are de nitions of terms:
A \standard" (represented by a number 1-8) is a speci c , well-de ned, physical device used
to determine systematic errors . F or example , standard 1 is a short in the 3.5 mm calibration
kit.
A standard \type" is one of ve basic types that de ne the form or structure of the model to
be used with that standard (short, open, load, delay/thru, and arbitrary impedance); standard
1 is of the type short in the 3.5 mm calibration kit.
Standard \coecients" are numerical characteristics of the standards used in the model
selected. F or example , the o set delay of the short is 32 ps in the 3.5 mm calibration kit.
A standard \class" is a grouping of one or more standards that determines which of the eight
standards are used at each step of the calibration. F or example , standard number 2 makes up
A re ection class .
Outline of Standard Modi cation
The following steps are used to modify or de ne user kit standard models , contained in the
analyzer memory . It is not possible to alter the built-in calibration kits; all modi cations will be
saved in the user kit.
de ning a new cal kit.
electrical characteristics .
NNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL
CAL KIT
SELECT CAL KIT .
NNNNNNNNNNNNNNNNNNN
RETURN
NNNNNNNNNNNNNNNNNNNN
MODIFY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DEFINE STANDARD .
. Refer to your
calibration kit manual for the numbers of the speci c standards in your kit. F or example ,
to select a short press
.
5-26 Optimizing Measurement Results

T able 5-3. Typical Calibration Kit Standard and Corresponding Number
in the previous step ,
SHORT should be the underlined softkey .
Do not press a softkey that is not underlined unless you want to change the
\type" of standard.
a.
NNNNNNNNNNNNN
OPEN
NNNNNNNN
C0 . Observe the value on the analyzer screen. Use the entry keys on the
analyzer front panel to change the value .
b.
NNNNNNNN
NNNNNNNN
Repeat the modi cation for
Ensure that the correct load type is underlined:
FIXED ,
NNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNN
OFFSET .
SPECIFY OFFSET
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
OFFSET DELAY and observe the value on the analyzer screen. T o
change the value , use the entry keys on the front panel.
OFFSET LOSS
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
OFFSET Z0
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MINIMUM FREQUENCY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MAXIMUM FREQUENCY
Ensure that the correct transmission line is underlined:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
WAVEGUIDE .
STD OFFSET DONE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
STD DONE (DEFINED) .
Optimizing Measurement Results 5-27

Saving the modi ed calibration constants
If you made modi cations to any of the standard de nitions , follow the remaining steps in this
procedure to assign a kit label, and store them in the non-volatile memory . The new set of
standard de nitions will be available under
USER KIT until you save another user kit.
NNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL
CAL KIT
MODIFY
LABEL KIT
ERASE TITLE . Use the front panel knob to
move the pointer to a character and press
SELECT LETTER .
T o enter titles , you may also use the optional external keyboard.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
KIT DONE (MODIFIED)
SAVE USER KIT .
Y ou may also save the user kit to disk, by selecting the particular kit at the
time you save a measurement result.
In order to use the TRL technique , the calibration standards characteristics must be entered
into the analyzer's user de ned calibration kit.
This example procedure shows you how to de ne a calibration kit to utilize a set of TRL
(THRU , REFLECT , LINE) standards . This example TRL kit contains the following:
\ ush" short for the REFLECT standard (0 second o set)
50 ohm transmission line with 80 ps of o set delay for the LINE
Hewlett-P ackard strongly recommends that you read product note 8510-8A
before you attempt to modify the standard de nitions . The part number of this
product note is 5091-3645E. Although the product note was written for the HP
8510 family of network analyzers , it also applies to the HP 8753D .
F or a discussion on TRL calibration, refer to \TRL/LRM Calibration" in
Chapter 6, \Application and Operation Concepts ."
Modify the Standard De nitions
NNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL
CAL KIT
MODIFY
DEFINE STANDARD
. (In this example the REFLECT standard is a SHORT .)
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SHORT
SPECIFY OFFSET
OFFSET DELAY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
STD OFFSET DONE
STD DONE (DEFINED)
5-28 Optimizing Measurement Results

DEFINE STANDARD
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DELAY/THRU
SPECIFY OFFSET
OFFSET DELAY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
STD OFFSET DONE
STD DONE (DEFINED)
DEFINE STANDARD
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DELAY/THRU
SPECIFY OFFSET
OFFSET DELAY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
STD OFFSET DONE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
F or the purposes of this example , change the name of the standard by pressing
and modifying the name to \LINE."
NNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DONE
STD DONE (DEFINED)
Assign the Standards to the V arious TRL Classes
NNNNNNNNNNNNNN
NNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SPECIFY CLASS
MORE
MORE
TRL REFLECT
TRL LINE OR MATCH
NNNNNNNNNNNNNNNNNNNNNNNNN
TRL THRU
SPECIFY CLASS DONE
Optimizing Measurement Results 5-29

T o enter the following label titles , an external keyboard may be used for
convenience .
NNNNNNNNNNNNNN
NNNNNNNNNNNNNN
MORE
MORE .
LABEL CLASS DONE .
Label the Calibration Kit
LABELKIT and create a label up to 8 characters long. F or this example , enter \TRL
NNNNNNNNNNNNNN
DONE .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
KIT DONE (MODIFIED)
SAVE USER KIT
In order to use the TRL technique , the calibration standards characteristics must be entered
into the analyzer's user de ned calibration kit.
This example procedure shows you how to de ne a calibration kit to utilize a set of TRM
(THRU , REFLECT , MA TCH) standards . This example TRM kit contains the following:
\ ush" short for the REFLECT standard (0 second o set)
50 ohm termination for the MA TCH (in nite length line)
Hewlett-P ackard strongly recommends that you read product note 8510-8A
before you attempt to modify the standard de nitions . The part number of this
product note is 5091-3645E. Although the product note was written for the
HP 8510 family of network analyzers , it also applies to the HP 8753D .
F or a discussion on TRL calibration, refer to \TRL/LRM Calibration" in
Chapter 6, \Application and Operation Concepts ."
Modify the Standard De nitions
N
NNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL
CAL KIT
MODIFY
DEFINE STANDARD
. (In this example the REFLECT standard is a SHORT .)
NNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SHORT
SPECIFY OFFSET
OFFSET DELAY
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
STD OFFSET DONE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
STD DONE (DEFINED)
5-30 Optimizing Measurement Results

DEFINE STANDARD
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DELAY/THRU
SPECIFY OFFSET
OFFSET DELAY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
STD OFFSET DONE
STD DONE (DEFINED)
DEFINE STANDARD
NNNNNNNNNNNNNN
LOAD
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
F or the purposes of this example , change the name of the standard by pressing
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ERASE TITLE , if a previous title exists , and then modify the name to \MA TCH".
DONE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
STD DONE (DEFINED)
Assign the Standards to the V arious TRM Classes
NNNNNNNNNNNNNN
NNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SPECIFY CLASS
MORE
MORE
TRL REFLECT
TRL LINE OR MATCH
TRL THRU
SPECIFY CLASS DONE
Optimizing Measurement Results 5-31

T o enter the following label titles , an external keyboard may be used for
convenience .
NNNNNNNNNNNNNN
NNNNNNNNNNNNNN
MORE
MORE .
LABEL CLASS DONE .
Label the Calibration Kit
LABELKIT and create a label up to 8 characters long. F or this example , enter \TRM
KIT1".
DONE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
KIT DONE (MODIFIED)
SAVE USER KIT
5-32 Optimizing Measurement Results
Label the Classes · Press

P ower Meter Measurement Calibration
Y ou can use the power meter to monitor and correct the analyzer source power to achieve
calibrated absolute power at the test port. Y ou can also use this calibration to set a reference
power for receiver power calibration, and mixer measurement calibration. The power meter
can measure and correct power in two ways:
continuous correction | each sweep mode
sample-and-sweep correction | single sweep mode
The time required to perform a power meter calibration depends on the source power ,
number of points tested, and number of readings taken. The parameters used to derive the
characteristic values in T able 5-4 are as follows:
T able 5-4.
Characteristic P ower Meter Calibration Sweep Speed and A ccuracy
Number of Readings Sweep Time
at T est P ort (dBm)
1 33 0.7
+5 2 64 0.2
3 95 0.1
1 48 0.7
15 2 92 0.2
3 123 0.1
1 194 0.7
30 2 360 0.2
3 447 0.1
1 Sweep speed applies to every sweep in continuous correction mode , and
to the rst sweep in sample-and-sweep mode . Subsequent sweeps in
sample-and-sweep mode will be much faster .
2 The accuracy values were derived by combining the accuracy of the
power meter and linearity of the analyzer's internal source , as well as
the mismatch uncertainty associated with the power sensor .
Loss of P ower Calibration Data
If your instrument state has not been saved after a power meter calibration,
the power correction data will be lost if any of the following circumstances
exists:
if you switch o the analyzer ac power and you haven't saved the correction
in an internal register
if you press
and you haven't saved the correction in an internal
if you change the sweep type (linear , log, list, CW , power) when the power
meter correction is activated
if you change the frequency when the sweep type is in log or list mode
Optimizing Measurement Results 5-33

Entering the P ower Sensor Calibration Data
Entering the power sensor calibration data compensates for the frequency response of the
power sensor , thus ensuring the accuracy of power meter calibration.
P eripherals" chapter for con guration procedures .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL
PWRMTR CAL
LOSS/SENSR LISTS
CAL FACTOR SENSOR A .
The analyzer shows the notation EMPTY , if you have not entered any segment information.
ADD
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FREQUENCY
.
CAL FACTOR and enter the correction factor that corresponds to the frequency that
you have entered in the previous step . Complete the correction factor entry by pressing
DONE .
Y ou may enter multiple segments in any order because the analyzer automatically sorts
them and lists them on the display by frequency value . The analyzer also automatically
interpolates the values between correction factor data points .
If you only enter one frequency segment the analyzer assumes that the single value is valid
over the entire frequency range of the correction.
After you have entered all the frequency segments , press
DONE .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
A ccess the \Segment Modify Menu" by pressing
PWRMTR CAL
LOSS/SENSR LISTS
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL FACTOR SENSOR A (or
CAL FACTOR SENSOR B , depending on where the segment is
that you want to edit).
Identify the segment that you want to edit by pressing
SEGMENT and using the
* and
keys to locate and position the segment next to the pointer (>), shown on the display . Or
SEGMENT and enter the segment number followed by
.
EDIT and then press either the
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL FACTOR key , depending of which
part of the segment you want to edit.
If you are modifying the frequency , enter the new value , followed by a
key .
If you are modifying the correction factor , enter the new value , followed by the
key .
DONE after you have nished modifying the segment.
If you want to edit any other segments , press
SEGMENT and follow the previous steps ,
starting with step 2.
5-34 Optimizing Measurement Results

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
A ccess the \Segment Modify Menu" by pressing
PWRMTR CAL
LOSS/SENSR LISTS
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL FACTOR SENSOR A (or
CAL FACTOR SENSOR B , depending on where the segment is
that you want to delete).
Identify the segment that you want to delete by pressing
SEGMENT and using the
* and
keys to locate and position the segment next to the pointer (>), shown on the display . Or
SEGMENT and enter the segment number followed by
.
DELETE .
The analyzer deletes the segment and moves the remainder of the segments up one number .
NNNNNNNNNNN
Y ou could also delete all the segments in a list by pressing
YES .
DONE when you are nished modifying the segment list.
Compensating for Directional Coupler Response
If you use a directional coupler to sample power in your measurement con guration, you
should enter the coupled arm power loss value into the power loss table , using the following
procedure . Y ou can enter the loss information in a single segment, and the analyzer will assume
that the value applies to the entire frequency range of the instrument. Or , you can input actual
measured power loss values at several frequencies using up to 55 segments , enhancing power
accuracy .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL
PWRMTR CAL
LOSS/SENSR LISTS
POWER LOSS .
The analyzer shows the notation EMPTY , if you have not entered any segment information.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
T o create the rst segment, press
FREQUENCY and enter a frequency of a correction
factor data point, followed by the appropriate key:
.
NNNNNNNNNNNNNN
LOSS and enter the power loss that corresponds to the attenuation of the directional
coupler (or power splitter) at the frequency that you have entered in the previous step .
Complete the power loss entry by pressing
DONE .
Remember to subtract the through arm loss from the coupler arm loss before
entering it into the power loss table , to ensure the correct power at the output
of the coupler .
required accuracy .
Y ou may enter multiple segments in any order because the analyzer automatically sorts them
and lists them on the display in increasing order of frequency .
If you only enter one frequency segment, the analyzer assumes that the single value is valid
over the entire frequency range of the correction.
NNNNNNNNNNNNN
After you have entered all the segments , press
DONE .
PWRMTR CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PWR LOSS ON to activate the power loss compensation.
Optimizing Measurement Results 5-35

Using Sample-and-Sweep Correction Mode
Y ou can use the sample-and-sweep mode to correct the analyzer output power and update
the power meter correction data table , during the initial measurement sweep . Because the
analyzer measures the actual power at each frequency point during the initial sweep , the initial
sweep time is signi cant. However , in this mode of operation the analyzer does not require the
power meter for subsequent sweeps . Therefore , this mode sweeps considerably faster than the
continuous correction mode .

Figure 5-8. Sample-and-Sweep Mode for P ower Meter Calibration
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SYSTEM CONTROLLER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SET ADDRESSES
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ADDRESS: P MTR/HPIB
Select the appropriate power meter by pressing
POWER MTR [ ] until the correct model
number is displayed (HP 436A or HP 438A/437).
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL
PWRMTR CAL and enter the test port power level that you want at the input to
your test device . F or example , if you enter
, the display will read CAL POWER 10 .
data point, press:
NUMBER OF READINGS
, (where n = the number of desired iterations).
If you increase the number of readings , the power meter correction time will substantially
increase .
5-36 Optimizing Measurement Results

PWRMTR CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ONE SWEEP
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TAKE CAL SWEEP .
Because power meter calibration requires a longer sweep time , you may want
to reduce the number of points before pressing
TAKE CAL SWEEP . After the
power meter calibration is nished, return the number of points to its original
value and the analyzer will automatically interpolate this calibration. Some
accuracy will be lost for the interpolated points .
The analyzer will use the data table for subsequent sweeps to correct the output power
level at each measurement point. Also , the status annunciator PC will appear on the
analyzer display .
Y ou can abort the calibration sweep by pressing
PWRMTR CAL OFF .
Using Continuous Correction Mode
Y ou can set the analyzer to update the correction table at each sweep (as in a leveling
application), using the continuous sample mode . When the analyzer is in this mode , it
continuously checks power at every point in each sweep . Y ou must keep the power meter
connected as shown in Figure 5-9. This mode is also known as power meter leveling, and the
speed is limited by the power meter .
Y ou may level at the input of a device under test, using a 2-resistor power
splitter or a directional coupler before the device; or level at the output of
the device using a 3-resistor power splitter or a bidirectional coupler after the
device .

Figure 5-9. Continuous Correction Mode for P ower Meter Calibration
device , as shown in Figure 5-9.
Optimizing Measurement Results 5-37

PWRMTR CAL and enter the test port power level that you want the analyzer
to maintain at the input to your test device . Compensate for the power loss of the power
splitter or directional coupler in the setup .
NUMBER OF READINGS
, (where n = the number of desired
iterations).
If you increase the number of readings , the power meter correction time will substantially
increase .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PWRMTR CAL
EACH SWEEP
TAKE CAL SWEEP to activate the power meter
correction.
T o Calibrate the Analyzer Receiver to Measure Absolute P ower
Y ou can use the power meter calibration as a reference to calibrate the analyzer receiver to
accurately measure absolute power . The following procedure shows you how to calibrate the
receiver to any power level.
NNNNNNNNNNNNNNNN
MENU
POWER (enter power level)
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL
PWRMTR CAL (enter power level)
TAKE CAL SWEEP
Because power meter calibration requires a longer sweep time , you may want
to reduce the number of points before pressing
TAKE CAL SWEEP . After the
power meter calibration is nished, return the number of points to its original
value and the analyzer will automatically interpolate this calibration.
The status notation PC will appear on the analyzer display . P ort 1 is now a calibrated source
of power .
NNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MEAS
INPUT PORTS
B
TEST PORT 1
This sets the source at PORT 1, and the measurement receiver to PORT 2, or input port B .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL
RECEIVER CAL (enter power level)
The receiver channel now measures power to a characteristic accuracy of 0.35 dB or better .
The accuracy depends on the match of the power meter , the source , and the receiver .
5-38 Optimizing Measurement Results

With this method, you use two precision matched adapters which are \equal." T o be equal,
the adapters must have the same match, Z
, insertion loss , and electrical delay . The adapters
in most HP calibration kits have matched electrical length, even if the physical lengths appear
di erent.

Figure 5-10. Calibrating for Noninsertable Devices
T o use this method, refer to Figure 5-10 and perform the following steps:
port.
The errors remaining after calibration with this method are equal to the di erences between
the two adapters that are used.
Optimizing Measurement Results 5-39
Matched A dapters

Modify the Cal Kit Thru De nition
With this method it is only necessary to use adapter B . The calibration kit thru de nition is
modi ed to compensate for the adapter and then saved as a user kit. However , the electrical
delay of the adapter must rst be found.
Measure the delay of the adapter by pressing
DELAY .
(see \De ne Standard Menus").
delay of adapter B .
Save this as a user kit.
5-40 Optimizing Measurement Results

Calibrating for Noninsertable Devices
A test device having the same sex connector on both the input and output cannot be connected
directly into a transmission test con guration. Therefore , the device is considered to be
noninsertable, and one of the following calibration methods must be performed:
modify the cal kit thru de nition

Figure 5-11. Noninsertable Device
Optimizing Measurement Results 5-41
adapter removal · matched adapters

The adapter removal technique provides a means to accurately measure noninsertable devices .
The following adapters are needed:
A dapter A1, which mates with port 1 of the device , must be installed on test set port 1.
A dapter A2, which mates with port 2 of the device , must be installed on test set port 2.
A dapter A3 must match the connectors on the test device . The e ects of this adapter will be
completely removed with this calibration technique .

Figure 5-12. A dapters Needed
The following requirements must also be met:
Calibration standards for performing a 2-port error correction for each connector type .
Speci ed electrical length of adapter A3 within 1/4 wavelength for the measurement
frequency range .
F or each port, a separate 2-port error correction needs to be performed to create two
calibration sets . The adapter removal algorithm uses the resultant data from the two
calibration sets and the nominal electrical length of the adapter to compute the adapter's
actual S-parameters . This data is then used to generate a separate third calibration set in
which the forward and reverse match and tracking terms are as if port 1 and port 2 could be
connected. This is possible because the actual S-parameters of the adapter are measured with
great accuracy , thus allowing the e ects of the adapter to be completely removed when the
third cal set is generated.
5-42 Optimizing Measurement Results
A dapter Removal

P erform the 2-port Error Corrections

Figure 5-13. Two-P ort Cal Set 1
connector type at port 1.
When using adapter removal calibration, you must save calibration sets to the
internal disk, not to internal memory .
Optimizing Measurement Results 5-43


Figure 5-14. Two-P ort Cal Set 2
connector type at port 2.
the Cal Kit Thru De nition."
Remove the A dapter
When the two sets of error correction les have been created (now referred to as \cal sets"),
the adapter may be removed.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL
MORE
ADAPTER REMOVAL . This brings up the following menu:
HELP ADAPT REMOVAL (This provides a quick reference guide to using the adapter
removal technique .)
RECALL CAL SETS
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ADAPTER DELAY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ADAPTER COAX
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ADAPTER WAVEGUIDE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
REMOVE ADAPTER
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RECALL CAL SETS to bring up the following two choices:
RECALL CAL PORT 1
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RECALL CAL PORT 2
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RECALL CAL SETS also brings up the internal (or external if internal not used) disk le
directory .
5-44 Optimizing Measurement Results
Press · · Press

In the following two steps , calibration data is recalled, not instrument states .
RECALL CAL PORT 1 .
RECALL CAL PORT 2 .
When complete , press
RETURN .
ADAPTER DELAY and enter the value .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Select the appropriate key:
ADAPTER WAVEGUIDE .
REMOVE ADAPTER to complete the technique for calculating the new error
coecients and overwrite the current active calibration set in use .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SELECT DISK
INTERNAL DISK
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
EXTERNAL DISK )
RETURN
SAVE STATE .

Figure 5-15. Calibrated Measurement
V erify the Results
Since the e ect of the adapter has been removed, it is easy to verify the accuracy of the
technique by simply measuring the adapter itself. Because the adapter was used during the
creation of the two cal sets , and the technique removes its e ects , measurement of the adapter
itself should show the S-parameters .
Optimizing Measurement Results 5-45

If unexpected phase variations are observed, this indicates that the electrical delay of the
adapter was not speci ed within a quarter wavelength over the frequency range of interest.
T o correct this , recall both cal sets , since the data was previously stored to disk, change the
adapter delay , and press
REMOVE ADAPTER .
5-46 Optimizing Measurement Results

The following is an example program for performing these same operations over HP-IB:
30 ! This demonstrates how to do adapter removal over HP--IB.
50 ASSIGN @Na TO 716
70 ! Select internal disk.
110 ! Assign file #1 to the filename that has a 2-port
120 ! cal previously performed for Port 1's connector.
160 ! Recall the cal set for Port 1.
200 ! Assign file #2 to the filename that has a 2-port
210 ! cal previously performed for Port 2's connector.
250 ! Recall the cal set for Port 2.
290 ! Set the adapter electrical delay.
330 ! Perform the "remove adapter" computation.
Optimizing Measurement Results 5-47

Making A ccurate Measurements of Electrically Long Devices
A device with a long electrical delay , such as a long length of cable or a SA W lter , presents
some unusual measurement problems to a network analyzer operating in swept frequency
mode . Often the measured response is dependent on the analyzer's sweep time , and incorrect
data may be obtained. At faster sweep rates , the magnitude of the response may seem to drop
and look distorted, while at slower sweep rates it looks correct. The results may indicate that a
cable has more loss than it truly does , or that a lter has some unusual ripple in the passband
which isn't really there .
This section describes the cause of this behavior , and how to accurately measure these
electrically long devices .
The Cause of Measurement Problems
When using a vector network analyzer to measure a device that has a long electrical delay
(T), the device's time delay causes a frequency shift between its input and output signals . The
frequency shift, F , equals the product of the sweep rate and the time delay:
F= dF/dt * T
Since frequency is changing with time as the analyzer sweeps , the time delay of the DUT
causes a frequency o set between its input and output. In the analyzer receiver , the test and
reference input signals will di er in frequency by F . Because the test signal frequency is
slightly di erent than the receiver frequency , the analyzer will err in measuring its magnitude
or phase . The faster the analyzer's sweep rate , the larger F becomes , and the larger the error
in the test channel.
The HP 8753D network analyzers do not sweep at a constant rate . The frequency range is
covered in several bands , and the sweep rate may be di erent in each band. So if an operator
sets up a broadband sweep with the minimum sweep time , the error in measuring a long device
will be di erent in each band, and the data will be discontinuous at each band edge . This can
produce confusing results which make it dicult to determine the true response of the device .
T o Improve Measurement Results
T o reduce the error in these measurements , the frequency shift, F , must be reduced. F can
be reduced by using the following three methods:
decreasing the sweep rate
decreasing the time delay (T)
Decreasing the Sweep Rate
The sweep rate can be decreased by increasing the analyzer's sweep time . T o increase the
analyzer's sweep time , press
MENU
SWEEP TIME [MANUAL] and use the front panel knob , the
keys , or the front panel keypad enter in the appropriate sweep time .
Selection of the appropriate sweep time depends on the device being measured; the longer the
electrical delay of the device under test, the slower the sweep rate must be . A good way to tell
when the sweep rate is slow enough is to put the vector network analyzer into a list frequency
mode of sweeping, and compare the data. In this mode , the vector network analyzer does not
sweep the frequency , but steps to each listed frequency point, stops , makes a measurement,
then goes on to the next point. Because errors do not occur in the list frequency mode , it can
be used to check the data. The disadvantage of the list frequency mode is that it is slower than
sweeping.
5-48 Optimizing Measurement Results

Decreasing the Time Delay
The other way to reduce F is by decreasing the time delay , T . Since T is a property of
the device that is being measured, it cannot literally be decreased. However , what can be
decreased is the di erence in delay times between the paths to the R channel and the B
channel. These times can be equalized by adding a length of cable to the R channel which has
approximately the same delay as the device under test.
This length of cable can be inserted between the R CHANNEL IN and OUT connectors on the
front panel of the analyzer . The delay of this cable must be less than 5s .
Optimizing Measurement Results 5-49

Y ou can increase the analyzer sweep speed by avoiding the use of some features that require
computational time for implementation and updating, such as bandwidth marker tracking.
Y ou can also increase the sweep speed by making adjustments to the measurement settings .
The following suggestions for increasing sweep speed are general rules that you should
experiment with:
decrease the frequency span
set the auto sweep time mode
widen the system bandwidth
reduce the averaging factor
reduce the number of measurement points
set the sweep type
use chop sweep mode
fast 2-port calibration mode
T o Decrease the Frequency Span
The hardware of the network analyzer sweeps the frequency range in separate bands , where
switching from band to band takes time . Modify the frequency span to eliminate as many band
switches as possible while maintaining measurement integrity . Refer to the following table to
identify the analyzer's band switch points:
T able 5-5. Band Switch P oints
0 .01 MHz to .3 MHz
1 .3 MHz to 3.3 MHz
2 3.3 MHz to 16 MHz
3 16 MHz to 31 MHz
4 31 MHz to 61 MHz
5 61 MHz to 121 MHz
6 121 MHz to 178 MHz
7 178 MHz to 296 MHz
8 296 MHz to 536 MHz
9 536 MHz to 893 MHz
10 893 MHz to 1.607 GHz
11 1.607 GHz to 3 GHz
12 (Option 006) 3 GHz to 4.95 GHz
13 (Option 006) 4.95 GHz to 6 GHz
5-50 Optimizing Measurement Results
Increasing Sweep Speed · use external calibration · Band Frequency Span

T o Set the A uto Sweep Time Mode
A uto sweep time mode is the default mode (the preset mode). This mode maintains the fastest
sweep speed possible for the current measurement settings .
MENU
SWEEP TIME
, to re-enter the auto mode .
Optimizing Measurement Results 5-51
Press · · 0

T o Widen the System Bandwidth
IF BW .
The following table shows the relative increase in sweep time as you decrease system
bandwidth. The characteristic values in the following table were derived using 201
measurement points .
IF B W Cycle Time (Seconds)
Full Span Narrow Sweep
3700 Hz .446 .150
3000 Hz .447 .176
1000 Hz .511 .312
300 Hz .944 .980
100 Hz 2.25 2.070
30 Hz 7.57 7.240
10 Hz 21.98 21.600
1 The listed sweep times correspond
to an HP 8753D analyzer being set
to a preset state for the full span,
and 2 GHz to 3 GHz for the narrow
span.
T o Reduce the A veraging F actor
By reducing the averaging factor (number of sweeps) or switching o averaging, you can
increase the analyzer's measurement speed. The time needed to compute averages can also
slow the sweep time slightly , in narrow spans .
A V G
AVG FACTOR .
.
If you want to switch o averaging, press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AVERAGING OFF .
T o Reduce the Number of Measurement P oints
MENU
NUMBER OF POINTS .
.
5-52 Optimizing Measurement Results

The analyzer sweep time does not change proportionally with the number of points , but as
indicated below .
Full Span Narrow Span
51 0.35 0.57 0.09 0.25
101 0.39 0.77 0.12 0.43
201 0.43 1.11 0.17 0.78
401 0.49 1.73 0.27 1.33
801 0.69 3.04 0.47 2.64
1601 1.09 5.7 0.87 5.30
1 The listed sweep times correspond to the
analyzer being set to a preset state , with a
6 GHz span. A 3 GHz span would have
faster sweep times .
T o Set the Sweep Type
Di erent sweep speeds are associated with the following three types of non-power sweeps .
Choose the sweep type that is most appropriate for your application.
MENU
SWEEP TYPE MENU .
LIN FREQ for the fastest sweep for a given number of xed points .
LIST FREQ for the fastest sweep when speci c non-linearly spaced frequency
points are of interest.
LOG FREQ for the fastest sweep when the frequency points of interest are in the
lower part of the frequency span selected.
T o View a Single Measurement Channel
Viewing a single channel will increase the measurement speed if the analyzer's channels are in
alternate , or uncoupled mode .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DISPLA Y
DUAL CHAN OFF .
CHAN 1 and
CHAN 2 to alternately view the two measurement channels .
If you must view both measurement channels simultaneously (with dual channel), use the
chop sweep mode , explained next.
Optimizing Measurement Results 5-53

T o A ctivate Chop Sweep Mode
Y ou can use the chop sweep mode to make two measurements at the same time . F or example ,
the analyzer can measure A/R and B/R simultaneously . Y ou can activate the chop mode by
or by the following the sequence below .
NNNNNNNNNNNNNN
MORE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CHOP A and B .
F or more information, refer to \Alternate and Chop Sweep Modes" in Chapter 6.
T o Use External Calibration
Ooading the error correction process to an external PC increases throughput on the network
analyzer . This can be accomplished with remote only commands . Refer to the HP 8753D
Programmer's Guide for information on how to use external calibration.
T o Use F ast 2-P ort Calibration
With the 2-port calibration on, faster measurements may be made by not measuring the reverse
path for every forward sweep . This is controlled by the test set switch command.
When making measurements using full two-port error-correction, the following types of test set
switching can be de ned by the user:
Hold: In this mode the analyzer does not switch between the test ports on every sweep . The
measurement stays on the active port after an initial cycling between the ports . The fastest
measurements can be made by using this type of test set switching. Pressing the
or changing to a di erent S-parameter measurement will cause the test set to switch and
cycle between the ports .
Continuous: In this mode the analyzer will switch between the test ports on every sweep .
Although this type of test set switching provides the greatest measurement accuracy , it also
takes the longest amount of time .
then the measurement stays on the active port for a user-de ned number of sweeps . After
the speci ed number of sweeps have been executed, the analyzer switches between the
test ports and begins the cycle again. This type of test set switching can provide improved
measurement accuracy over the hold mode and faster measurement speeds than continuous
mode .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL
MORE
TEST SET SW
The analyzer will then display
TEST SET SW HOLD .
The analyzer will then display
TEST SET SW CONTINUOUS .
5-54 Optimizing Measurement Results

The analyzer will then display
TEST SET SW X SWEEPS .
Optimizing Measurement Results 5-55
· ·

Dynamic range is the di erence between the analyzer's maximum allowable input level and
minimum measurable power . F or a measurement to be valid, input signals must be within these
boundaries . The dynamic range is a ected by these factors:
test port input power
test port noise oor
T o Increase the T est P ort Input P ower
Y ou can increase the analyzer's source output power so that the test device output power is at
the top of the measurement range of the analyzer test port.
POWER and enter the new source power level, followed by
.
TEST PORT INPUT D AMA GE LEVEL: +26 dBm
T o Reduce the Receiver Noise Floor
Since the dynamic range is the di erence between the analyzer's input level and its noise oor ,
using the following techniques to lower the noise oor will increase the analyzer's dynamic
range .
Each tenfold reduction in IF (receiver) bandwidth lowers the noise oor by 10 dB . F or example ,
changing the IF bandwidth from 3 kHz to 300 Hz, you will lower the noise oor by about
10 dB .
IF BW .
.
Changing Measurement A veraging
Y ou can apply weighted averaging of successive measurement traces to remove the e ects of
random noise .
AVERAGING FACTOR .
.
AVERAGING ON .
Refer to the \Application and Operation Concepts" chapter for more information on
averaging.
5-56 Optimizing Measurement Results

Y ou can use two analyzer functions to help reduce the e ect of noise on the data trace:
T o A ctivate A veraging
The noise is reduced with each new sweep as the e ective averaging factor increments .
A VG
AVERAGING FACTOR .
.
AVERAGING ON .
Refer to the \Application and Operation Concepts" chapter for more information on
averaging.
T o Change System Bandwidth
By reducing the system bandwidth, you reduce the noise that is measured during the sweep .
While averaging requires multiple sweeps to reduce noise , narrowing the system bandwidth
reduces the noise on each sweep .
A VG
IF BW .
.
Narrower system bandwidths cause longer sweep times . When in auto sweep time mode , the
analyzer uses the fastest sweep time possible for any selected system bandwidth. A uto sweep
time mode is the default (preset) analyzer setting.
T o reduce receiver crosstalk you can do the following:
perform a response and isolation measurement calibration
set the sweep to the alternate mode
Alternate sweep is intended for measuring wide dynamic range devices , such as high pass and
bandpass lters . This sweep mode removes a type of leakage term through the device under
test, from one channel to another .
T o set the alternate sweep , press
MORE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ALTERNATE A AND B .
Refer to the procedures , located earlier in this chapter for a response and isolation
measurement calibration procedure .
Optimizing Measurement Results 5-57

T o reduce time during recall and frequency changes , the raw o set function and the spur
avoidance function can be turned o . T o turn these functions o , press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SYSTEM
CONFIGURE MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RAW OFFSET OFF
SPUR AVOID OFF
The raw o set function is normally on and controls the sampler and attenuator o sets . The
spur avoidance function is normally on and generates values as part of the sampler o set table .
The creation of this table takes considerable time during a recall of an instrument state .
T o save time at recalls and during frequency changes , both functions should be turned o . This
will avoid generating the sampler o set table .
Raw o sets may be turned on or o individually for each channel. They follow the channel
coupling. F or dual channel operation, raw o sets should be turned o for each channel if the
channels are uncoupled. Spur avoidance is always coupled between channels , therefore both
channels are turned on or o at the same time .
Both functions must be turned o to realize the recall time savings .
The following table lists the recall state times with the raw o sets and spur avoidance
functions on or o .
T able 5-6. Typical Recall State Times
Operations Channel P oints Raw O set
Recall and Sweep Dual Chan. 1601 On 3.89 3.18
Recall and Sweep Dual Chan. 1601 O 2.008 1.298
Dual Chan. 1601 n/a 0.71 no recall
Recall and Sweep Dual Chan. 201 On 0.955 .740
Recall and Sweep Dual Chan. 201 O 0.734 .519
Dual Chan. 201 n/a 0.215 no recall
Recall and Sweep Single Chan. 1601 On 2.134 1.424
Recall and Sweep Single Chan. 1601 O 1.251 .541
Recall;OPC?;SING; or , for sweep only , OPC?;SING;.
5-58 Optimizing Measurement Results

Understanding Spur A voidance
In the 400 MHz to 3 GHz range , where the source signal is created by heterodyning two higher
frequency oscillators , unwanted spurious mixing products from the source may be present at
the output. These spurs can become apparent in lter measurements when lters have greater
than 80 dB rejection.
Spur avoidance slightly moves the frequency of both oscillators such that the source frequency
remains the same but the spurious mixing products shift out of the measurement receiver
range . The calculation of the exact frequency points where the shifting must occur (stored
in the sampler o set table) increases the time needed to change or recall instrument states .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SPUR AVOID OFF and
RAW OFFSET OFF eliminates this calculation.
Optimizing Measurement Results 5-59
Selecting



Application and Operation Concepts
This chapter provides conceptual information on the following primary operations and
applications that are achievable with the HP 8753D network analyzer .
HP 8753D System operation
A ctive channel keys
Where to Look for More Information
A dditional information about many of the topics discussed in this chapter is located in the
following areas:
Chapter 2, \Making Measurements ," contains step-by-step procedures for making
measurements or using particular functions .
Chapter 3, \Mixer Measurements ," contains step-by-step procedures for making
measurements of mixers .
Chapter 5, \Optimizing Measurement Results ," describes techniques and functions for
achieving the best measurement results .
Chapter 7, \Speci cations and Measurement Uncertainties ," de nes the performance
capabilities of the analyzer .
Chapter 8, \Menu Maps ," shows softkey menu relationships .
Chapter 9, \Key De nitions ," describes all the front panel keys and softkeys .
Application and Operation Concepts 6-1

HP 8753D System Operation
Network analyzers measure the re ection and transmission characteristics of devices and
networks . A network analyzer test system consists of the following:
The analyzer applies a signal that is transmitted through the test device , or re ected from its
input, and then compares it with the incident signal generated by the swept RF source . The
signals are then applied to a receiver for measurement, signal processing, and display .
The HP 8753D vector network analyzer integrates a high resolution synthesized RF source ,
test set, and a dual channel three-input receiver to measure and display magnitude , phase ,
and group delay of transmitted and re ected power . With Option 010, the analyzer has the
additional capability of transforming measured data from the frequency domain to the time
domain. Other options are explained in Chapter 1, \HP 8753D Description and Options ."
Figure 6-1 is a simpli ed block diagram of the network analyzer system. A detailed block
diagram of the analyzer is provided in the HP 8753D Network Analyzer Service Guide together
with a theory of system operation.

Figure 6-1. Simpli ed Block Diagram of the Network Analyzer System
The Built-In Synthesized Source
The analyzer's built-in synthesized source produces a swept RF signal or CW (continuous wave)
signal in the range of 30 kHz to 3.0 GHz. The HP 8753D Option 006 is able to generate signals
up to 6 GHz. The RF output power is leveled by an internal ALC (automatic leveling control)
circuit. T o achieve frequency accuracy and phase measuring capability , the analyzer is phase
locked to a highly stable crystal oscillator . F or this purpose , a portion of the transmitted signal
is routed to the R channel input of the receiver , where it is sampled by the phase detection
loop and fed back to the source . Some portion of the RF source signal must always be sent to
the R channel input. The level must be between 0 dB and -35 dBm.
The Source Step Attenuator
The step attenuator contained in the source is used to adjust the power level to the test device
without changing the level of the incident power in the reference path.
6-2 Application and Operation Concepts

The Built-In T est Set
The HP 8753D features a built-in test set that provides connections to the test device , as well
as to the signal-separation devices . The signal separation devices are needed to separate the
incident signal from the transmitted and re ected signals . The incident signal is applied to the
R channel input through a jumper cable on the front panel. Meanwhile , the transmitted and
re ected signals are internally routed from the test port couplers to the inputs of the A and B
sampler/mixers in the receiver . P ort 1 is connected to the A input and port 2 is connected to
the B input.
The test set contains the hardware required to make simultaneous transmission and re ection
measurements in both the forward and reverse directions . An RF path switch in the test set
allows reverse measurements to be made without changing the connections to the test device .
The receiver block contains three sampler/mixers for the R, A, and B inputs . The signals
are sampled, and mixed to produce a 4 kHz IF (intermediate frequency). A multiplexer
sequentially directs each of the three signals to the ADC (analog to digital converter)
where it is converted from an analog to a digital signal. The signals are then measured and
processed for viewing on the display . Both amplitude and phase information are measured
simultaneously , regardless of what is displayed on the analyzer .
A microprocessor takes the raw data and performs all the required error correction, trace
math, formatting, scaling, averaging, and marker operations , according to the instructions
from the front panel or over HP-IB . The formatted data is then displayed. The data processing
sequence is described in \Data Processing" later in this chapter .
Required P eripheral Equipment
Measurements require calibration standards for vector accuracy enhancement (error-
correction), and cables for interconnections . Model numbers and details of compatible power
splitters , calibration kits , and cables are provided in Chapter 11, \Compatible P eripherals ."
Application and Operation Concepts 6-3
The Receiver Block · The Microprocessor

The analyzer's receiver converts the R, A, and B input signals into useful measurement
information. This conversion occurs in two main steps:
The swept high frequency input signals are translated to xed low frequency IF signals , using
analog sampling or mixing techniques . (Refer to the HP 8753D Network Analyzer Service
Guide for more details on the theory of operation.)
The IF signals are converted into digital data by an analog to digital converter (ADC). From
this point on, all further signal processing is performed mathematically by the analyzer
microprocessors .
The following paragraphs describe the sequence of math operations and the resulting data
arrays as the information ows from the ADC to the display . They provide a good foundation
for understanding most of the response functions , and the order in which they are performed.
Figure 6-2 is a data processing ow diagram that represents the ow of numerical data from IF
detection to display . The data passes through several math operations , denoted in the gure by
single line boxes . Most of these operations can be selected and controlled with the front panel
response block menus . The data, stored in arrays along the way and denoted by double line
boxes , are places in the ow path where data is accessible via HP-IB .

Figure 6-2. Data Processing Flow Diagram
6-4 Application and Operation Concepts
Data Processing

While only a single ow path is shown, two identical paths are available , corresponding to
channel 1 and channel 2. When the channels are uncoupled, each channel is processed and
controlled independently .
measurement at a single stimulus value . Most data processing operations are performed
point-by-point; some involve more than one point.
a sequence of stimulus values . A few data processing operations require that a full sweep
of data is available . The number of points per sweep can be de ned by the user . The units
of the stimulus values (such as power , frequency , and time) can change , depending on the
sweep mode , although this does not generally a ect the data processing path.
The ADC (analog-to-digital converter) converts the R, A, and B inputs (already down-converted
to a xed low frequency IF) into digital words . (The A UX INPUT connector on the rear panel is
a fourth input.) The ADC switches rapidly between these inputs , so they are converted nearly
simultaneously .
This detection occurs in the digital lter , which performs the discrete F ourier transform
(DFT) on the digital words . The samples are converted into complex number pairs (real plus
imaginary , R+jX). The complex numbers represent both the magnitude and phase of the IF
signal. If the A UX INPUT is selected, the imaginary part of the pair is set to zero . The DFT
lter shape can be altered by changing the IF bandwidth, which is a highly e ective technique
for noise reduction.
These calculations are performed if the selected measurement is a ratio of two inputs (for
example , A/R or B/R). This is a complex divide operation. If the selected measurement is
absolute (such as A or B), no calculations are performed. The R, A, and B values are also split
into channel data at this point.
The next digital processing technique used is sampler/IF correction. This process digitally
corrects for frequency response errors (both magnitude and phase , primarily sampler rollo ) in
the analog down-conversion path.
Sweep-T o-Sweep A veraging
A veraging is another noise reduction technique . This calculation involves taking the complex
exponential average of several consecutive sweeps . This technique cannot be used with
single-input measurements .
Application and Operation Concepts 6-5

These data arrays store the results of all the preceding data processing operations . (Up to
this point, all processing is performed real-time with the sweep by the IF processor . The
remaining operations are not necessarily synchronized with the sweep , and are performed
by the main processor .) When full 2-port error correction is on, the raw arrays contain all
four S-parameter measurements required for accuracy enhancement. When the channels are
COUPLED CH OFF ), there may be as many as eight raw arrays . These arrays are
directly accessible via HP-IB . Notice that the numbers here are still complex pairs .
Raw arrays contain the pre-raw data which has sampler and attenuator o set applied.
V ector Error-correction (A ccuracy Enhancement)
Error-correction is performed next, if a measurement calibration has been performed and
correction is activated. Error-correction removes repeatable systematic errors (stored in the
error coecient arrays) from the raw arrays . This can vary from simple vector normalization to
full 12-term error-correction.
The results of error-correction are stored in the data arrays as complex number pairs . These
are subsequently used whenever correction is on, and are accessible via HP-IB .
If the data-to-memory operation is performed, the data arrays are copied into the memory
arrays .
This operation selects either the data array , memory array , or both to continue owing through
the data processing path. In addition, the complex ratio of the two (data/memory) or the
di erence (datamemory) can also be selected. If memory is displayed, the data from the
memory arrays goes through exactly the same processing ow path as the data from the data
arrays .
Gating (Option 010 Only)
This digital ltering operation is associated with time domain transformation. Its purpose
is to mathematically remove unwanted responses isolated in time . In the time domain, this
can be viewed as a time-selective bandpass or bandstop lter . (If both data and memory are
displayed, gating is applied to the memory trace only if gating was on when data was stored
into memory .)
The Electrical Delay Block
This block involves adding or subtracting phase in proportion to frequency . This is equivalent
to \line-stretching" or arti cially moving the measurement reference plane . This block also
includes the e ects of port extensions as well as electrical delay .
This converts the measured S-parameter data to the equivalent complex impedance (Z) or
admittance (Y) values , or to inverse S-parameters (1/S).
Transform (Option 010 Only)
This transform converts frequency domain information into the time domain when it is
activated . The results resemble time domain re ectometry (TDR) or impulse-response
measurements . The transform uses the chirp-Z inverse fast F ourier transform (FFT) algorithm
to accomplish the conversion. The windowing operation, if enabled, is performed on the
6-6 Application and Operation Concepts

frequency domain data just before the transform. (A special transform mode is available
to \demodulate" CW sweep data, with time as the stimulus parameter , and display spectral
information with frequency as the stimulus parameter .)
This operation converts the complex number pairs into a scalar representation for display ,
according to the selected format. This includes group delay calculations . These formats are
often easier to interpret than the complex number representation. (P olar and Smith chart
formats are not a ected by the scalar formatting.) Notice in Figure 6-2 that after formatting, it
is impossible to recover the complex data.
This noise reduction technique smoothes noise on the trace . Smoothing is also used to set the
aperture for group delay measurements .
When smoothing is on, each point in a sweep is replaced by the moving average value of
several adjacent (formatted) points . The number of points included depends on the smoothing
aperture , which can be selected by the user . The e ect is similar to video ltering. If data and
memory are displayed, smoothing is performed on the memory trace only if smoothing was on
when data was stored into memory .
The data processing results are now stored in the format arrays . Notice in Figure 6-2 that the
marker values and marker functions are all derived from the format arrays . Limit testing is
also performed on the formatted data. The format arrays are accessible via HP-IB .
O set and Scale
These operations prepare the formatted data for display . This is where the reference line
position, reference line value , and scale calculations are performed, as appropriate to the
format.
The display memory stores the display image for presentation on the analyzer . The information
stored includes graticules , annotation, and softkey labels . If user display graphics are written,
these are also stored in display memory . When a print or plot is made , the information is taken
from display memory .
The display is updated frequently and synchronously with the data processing operations .
Application and Operation Concepts 6-7

A ctive Channel K eys
The analyzer has two digital channels for independent measurements . Two di erent sets
of data can be measured simultaneously , for example , the re ection and transmission
characteristics of a device , or one measurement with two di erent frequency spans . The
analyzer can separately , or simultaneously , show the data.

Figure 6-3. A ctive Channel K eys
CHAN 1 and
CHAN 2 keys shown in Figure 6-3 allow you to select the \active
channel." The front panel keys currently allow you to control the active channel. All of the
channel-speci c functions that you select apply to the active channel. The current active
channel is indicated by an amber LED adjacent to the corresponding channel key .
The analyzer has dual channel capability , so that you can view both the active and inactive
channel traces , either overlaid or on separate graticules one above the other (split display).
The dual channel and split display features are available in the display menus . Refer to
\Display Menu" later in this chapter for illustrations and descriptions of the di erent display
capabilities .
Uncoupling Stimulus V alues Between Channels
Y ou can uncouple the stimulus values between the two display channels by pressing
COUPLED CH OFF . This allows you to assign di erent stimulus values for each channel; it's
almost like having the use of a second analyzer . The coupling and uncoupling of the stimulus
values for the two channels are independent of the display and marker functions . Refer to
\Channel Stimulus Coupling" later in this chapter for a listing of the stimulus parameters
associated with the coupled channel mode .
Measurement markers can have the same stimulus values (coupled) for the two channels , or
they can be uncoupled for independent control in each channel. Refer to \Markers" later in
this chapter for more information about markers .
6-8 Application and Operation Concepts
The · Dual Channel · Coupled Markers

Entry Block K eys
The entry block, illustrated in Figure 6-4, includes the numeric and units keypad, the knob ,
and the step keys . Y ou can use these in combination with other front panel keys and softkeys
for the following reasons:
to modify the active entry
to enter or change numeric data
to change the value of the active marker
Generally , the keypad, knob , and step keys can be used interchangeably .
Before you can modify a function, you must activate the particular function by pressing the
corresponding front panel key or softkey . Then you can modify the value directly with the
knob , the step keys , or the digits keys and a terminator .
If no other functions are activated, the knob moves the active marker .

Figure 6-4. Entry Block
The units terminator keys are the four keys in the right column of the keypad. Y ou must
use these keys to specify units of numerical entries from the keypad. A numerical entry is
incomplete until a terminator is supplied. The analyzer indicates that an input is incomplete
by a data entry arrow pointing at the last entered digit in the active entry area. When you
press the units terminator key , the arrow is replaced by the units you selected. The units are
abbreviated on the terminator keys as follows:
= basic units: dB , dBm, degrees , seconds , Hz, or dB/GHz (may be used to terminate
unitless entries such as averaging factor)
Application and Operation Concepts 6-9

Y ou can use the knob to make continuous adjustments to current measurement parameter
values or the active marker position. V alues changed by the knob are e ective immediately ,
and require no units terminator .
Y ou can use the step keys
* (up) and
(down) to step the current value of the active
function up or down. The analyzer de nes the steps for di erent functions . No units
terminator is required. F or editing a test sequence , you can use these keys to scroll through the
displayed sequence .
Y ou can use this key to clear and turn o the active entry area, as well as any displayed
prompts , error messages , or warnings . Use this function to clear the display before printing or
plotting. This key also helps prevent changing active values accidentally by moving the knob .
Y ou can use this key to delete the last entry , or the last digit entered from the numeric keypad.
Y ou can also use this key in one of two ways for modifying a test sequence:
deleting a single-key command that you may have pressed by mistake (for example
Trans:FWD S21 (B/R) )
deleting the last digit in a series of digits that you may have input, as long as you haven't yet
pressed a terminator (for example if you pressed
12 but did not press
Y ou can use this key to add a decimal point to the number you entered.
Y ou can use this key to add a minus sign to the number you entered.
6-10 Application and Operation Concepts


Figure 6-5. Stimulus Function Block
The stimulus function block keys are used to de ne the source RF output signal to the test
device by providing control of the following parameters:
time domain start and stop times (Option 010 Only)
power sweep start and stop values
RF power level and power ranges
number of data points
channel and test port coupling
De ning Ranges with Stimulus K eys
STOP
CENTER
, and
keys are used to de ne the swept frequency range ,
time domain range (Option 010), or power sweep range . The range can be expressed as either
start/stop or center/span. When one of these keys is pressed, its function becomes the active
function. The value is displayed in the active entry area and can be changed with the knob ,
step keys , or numeric keypad. Current stimulus values for the active channel are also displayed
along the bottom of the graticule . Frequency values can be blanked for security purposes ,
using the display menus .
The preset stimulus mode is frequency , and the start and stop stimulus values are set to 30
kHz and 3 GHz (or 6 GHz with Option 006) respectively . In the time domain (Option 010) or in
CW time mode , the stimulus keys refer to time (with certain exceptions). In power sweep , the
stimulus value is in dBm.
Because the display channels are independent, the stimulus signals for the two channels can be
uncoupled and their values set independently . The values are then displayed separately if the
instrument is in dual channel display mode . In the uncoupled mode with dual channel display
the instrument takes alternate sweeps to measure the two sets of data. Channel stimulus
Application and Operation Concepts 6-11

coupling is explained in the \Stimulus Menu" section, and dual channel display capabilities are
explained in the \Display Menu" section located later in this chapter .
key provides access to the stimulus menu, which consists of softkeys that activate
stimulus functions or provide access to additional menus . These softkeys are used to de ne and
control all stimulus functions other than start, stop , center , and span. The following softkeys
are located within the stimulus menu:
POWER provides access to the power menu.
SWEEP TIME allows you to specify the sweep time .
TRIGGER MENU provides access to the trigger menu.
NUMBER of POINTS allows you to specify the number of measurement points per sweep .
MEASURE RESTART allows you to cause the current measurement to abort and a new
measurement to begin. With two-port error-correction activated, pressing this softkey causes
all four S-parameters to be remeasured.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
COUPLED CH ON off allows you to couple or uncouple the stimulus functions of the two
display channels .
CW FREQ allows you to specify the CW frequency .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SWEEP TYPE MENU provides access to the sweep type menu.
6-12 Application and Operation Concepts
Stimulus Menu · The

The P ower Menu
The power menu is used to de ne and control analyzer power . It consists of the following
softkeys:
PWR RANGE AUTO man allows you to select power ranges automatically or manually .
POWER RANGES provides access to the power range menu.
SLOPE compensates for power loss versus the frequency sweep , by sloping the output power
upwards proportionally to frequency .
SLOPE on OFF toggles the power slope function on or o .
SOURCE PWR ON off allows you to switch the source power on or o . When a power trip
occurs , the trip is reset by selecting
SOURCE PWR ON .
CHAN PWR [COUPLED] allows you to couple or uncouple channel power .
PORT POWER allows you to couple or uncouple port power .
Understanding the P ower Ranges
The built-in synthesized source contains a programmable step attenuator that allows you to
directly and accurately set power levels in eight di erent power ranges . Each range has a total
span of 25 dB . The eight ranges cover the instrument's full operating range from +10 dBm to
85 dBm (see Figure 6-6). A power range can be selected either manually or automatically .
PWR RANGE AUTO , you can enter any power level within the total operating range
of the instrument and the source attenuator will automatically switch to the corresponding
range .
Each range overlaps its adjacent ranges by 15 dB , therefore , certain power levels are
designated to cause the attenuator to switch to the next range so that optimum (leveled)
performance is maintained. These transition points exist at 10 dB from the top of a range and
at +5 dB from the bottom of a range . This leaves 10 dB of operating range . By turning the
PORT POWER being the active function, you can hear the attenuator switch as
these transitions occur (see Figure 6-6).
PWR RANGE MAN , you must rst enter the power ranges menu and manually select
the power range that corresponds to the power level you want to use . This is accomplished by
POWER RANGES softkey and then selecting one of the twelve available ranges . In
this mode , you will not be able to use the step keys , front panel knob , or keypad entry to select
power levels outside the range limits . This feature is necessary to maintain accuracy once a
measurement calibration is activated.
When a calibration is active , the power range selection is switched from auto to manual mode ,
and PRm appears on the display .
Application and Operation Concepts 6-13

After measurement calibration, you can change the power within a range
and still maintain nearly full accuracy . In some cases better accuracy can
be achieved by changing the power within a range . It can be useful to set
di erent power levels for calibration and measurement to minimize the e ects
of sampler compression or noise oor .
If you decide to switch power ranges , the calibration is no longer valid and
accuracy is no longer speci ed. However , the analyzer leaves the correction on
even though it's invalid.
The annotation C? will be displayed whenever you change the power after
calibration.

Figure 6-6. P ower Range Transitions in the A utomatic Mode
6-14 Application and Operation Concepts

P ower Coupling Options
There are two methods you can use to couple and uncouple power levels with the HP 8753D:
By uncoupling the channel powers , you e ectively have two separate sources . Uncoupling the
test ports allows you to have di erent power levels on each port.
CHAN POWER [COUPLED] toggles between coupled and uncoupled channel power . With the
channel power coupled, the power levels are the same on each channel. With the channel
power uncoupled, you can set di erent power levels for each channel. F or the channel
power to be uncoupled, the other channel stimulus functions must also be uncoupled
COUPLED CH OFF ).
T est port coupling
PORT POWER [COUPLED] toggles between coupled and uncoupled test ports . With the test
ports coupled, the power level is the same at each port. With the ports uncoupled, you can
set a di erent power level at each port. This can be useful, for example , if you want to
simultaneously perform a gain and reverse isolation measurement on a high-gain ampli er using
the dual channel mode to display the results . In this case , you would want the power in the
) much lower than the power in the reverse direction (S
).
Application and Operation Concepts 6-15

SWEEP TIME softkey selects sweep time as the active entry and shows whether the
automatic or manual mode is active . The following explains the di erence between automatic
and manual sweep time:
Manual sweep time . As long as the selected sweep speed is within the capability of the
instrument, it will remain xed, regardless of changes to other measurement parameters . If
you change measurement parameters such that the instrument can no longer maintain the
selected sweep time , the analyzer will change to the fastest sweep time possible .
A uto sweep time . A uto sweep time continuously maintains the fastest sweep speed possible
with the selected measurement parameters .
Sweep time refers only to the time that the instrument is sweeping and taking data, and
does not include the time required for internal processing of the data, retrace time , or
bandswitching time . A sweep speed indicator " is displayed on the trace for sweep times longer
than 1.0 second. F or sweep times faster than 1.0 second the " indicator appears in the status
Manual Sweep Time Mode
When this mode is active , the softkey label reads
SWEEP TIME [MANUAL] . This mode is
engaged whenever you enter a sweep time greater than zero . This mode allows you to select a
xed sweep time . If you change the measurement parameters such that the current sweep time
is no longer possible , the analyzer will automatically increase to the next fastest sweep time
possible . If the measurement parameters are changed such that a faster sweep time is possible ,
the analyzer will not alter the sweep time while in this mode .
A uto Sweep Time Mode
When this mode is active , the softkey label reads
SWEEP TIME [AUTO] . This mode is engaged
whenever you enter
as a sweep time . A uto sweep time continuously maintains the
fastest sweep time possible with the selected measurement parameters .
The minimum sweep time is dependent on the following measurement parameters:
the number of points selected
sweep-to-sweep averaging in dual channel display mode
In addition to the parameters listed above , the actual cycle time of the analyzer is also
dependent on the following measurement parameters:
6-16 Application and Operation Concepts

time domain (Option 010 Only)
Use T able 6-1 to determine the minimum cycle time for the listed measurement parameters .
The values listed represent the minimum time required for a CW time measurement with
averaging o .
T able 6-1. Minimum Cycle Time (in seconds)
3700 Hz 3000 Hz 1000 Hz 300 Hz 100 Hz 30 Hz 10 Hz
11 0.0041 s 0.0055 s 0.012 s 0.037 s 0.108 s 0.359 s 1.14 s
51 0.0191 s 0.0255 s 0.060 s 0.172 s 0.504 s 1.660 s 5.30 s
101 0.0379 s 0.0505 s 0.120 s 0.341 s 0.998 s 3.300 s 10.5 s
201 0.0754 s 0.1005 s 0.239 s 0.679 s 1.990 s 6.600 s 20.9 s
401 0.1504 s 0.2005 s 0.476 s 1.355 s 3.960 s 13.10 s 41.7 s
801 0.3004 s 0.4005 s 0.951 s 2.701 s 7.910 s 26.10 s 83.3 s
1601 0.6004 s 0.8005 s 1.901 s 5.411 s 15.80 s 52.20 s 166.5 s
Application and Operation Concepts 6-17
Number of · IF Bandwidth · P oints

The trigger menu is used to select the type and number of groups for the sweep trigger . The
following is a description of the softkeys located within this menu:
HOLD freezes the data trace on the display , and the analyzer stops sweeping and taking data.
The notation \Hld" is displayed at the left of the graticule . If the " indicator is on at the left
side of the display , trigger a new sweep with
SINGLE .
SINGLE takes one sweep of data and returns to the hold mode .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NUMBER of GROUPS triggers a user-speci ed number of sweeps , and returns to the hold
mode . This function can be used to override the test set hold mode (indicated by the
switch (Option 007) and attenuator are not protected against unwanted continuous switching
. This occurs in a full two-port calibration, in a measurement of two di erent parameters
that require power out from both ports , or when the channels are uncoupled and a di erent
power level is set for each channel.
If averaging is on, the number of groups should be at least equal to the averaging factor
selected, to allow measurement of a fully averaged trace . Entering a number of groups resets
the averaging counter to 1.
CONTINUOUS is the standard sweep mode . The sweep is triggered automatically and
continuously and the trace is updated with each sweep .
TRIGGER: TRIG OFF switches o external trigger mode .
EXT TRIG ON SWEEP is used when the sweep is triggered on an externally generated signal
that is connected to the rear panel EXT TRIGGER input. The sweep is started with a high
to low transition of a TTL signal. If this key is pressed when no external trigger signal is
connected, the notation ''Ext'' is displayed at the left side of the display to indicate that
the analyzer is waiting for a trigger . When a trigger signal is connected, the ''Ext'' notation
is replaced by the sweep speed indicator either in the status notations area or on the trace .
External trigger mode is allowed in every sweep mode .
EXT TRIG ON POINT is similar to the trigger on sweep , but triggers on each data point in a
sweep .
MANUAL TRG ON POINT waits for a manual trigger for each point. Subsequent pressing of
this softkey triggers each measurement. The annotation ''man'' appears at the left side of
the display when the instrument is waiting for the trigger to occur . This feature is useful in a
test sequence when an external device or instrument requires changes at each point.
6-18 Application and Operation Concepts
Trigger Menu

Source Attenuator Switch Protection
The programmable step attenuator of the source can be switched between port 1 and port 2
when the test port power is uncoupled or between channel 1 and channel 2 when the channel
power is uncoupled. T o avoid premature wear of the attenuator , measurement con gurations
requiring continuous switching between di erent power ranges are not allowed.
F or example , channels 1 and 2 of the analyzer are decoupled, power levels in two di erent
ranges are selected for each channel, and dual channel display is engaged. T o prevent
continuous switching between the two power ranges , the analyzer automatically engages
the test set hold mode after measuring both channels once . The active channel continues
to be updated each sweep while the inactive channel is placed in the hold mode . (The
status annotation tsH appears on the left side of the display .) If averaging is on, the test
set hold mode does not engage until the speci ed number of sweeps is completed. The
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NUMBER OF GROUPS (see \Trigger Menu") softkeys can override this
protection feature .
Allowing Repetitive Switching of the Attenuator
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NUMBER OF GROUPS (see \Trigger Menu") softkeys allow
measurements which demand repetitive switching of the step attenuator . Use these softkeys
with caution; repetitive switching can cause premature wearing of the attenuator .
MEASURE RESTART causes one measurement to occur before activating the test set hold
mode .
NUMBER OF GROUPS (see \Trigger Menu") causes a speci ed number of measurements to
occur before activating the test set hold mode .
Application and Operation Concepts 6-19
MEASURE RESTART and · The · MEASURE RESTART and

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
COUPLED CH on OFF toggles the channel coupling of stimulus values . With
(the preset condition), both channels have the same stimulus values . (The inactive channel
takes on the stimulus values of the active channel.)
In the stimulus coupled mode , the following parameters are coupled:
Coupling of stimulus values for the two channels is independent of
DUAL CHAN on OFF in
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
the display menu and
MARKERS: UNCOUPLED in the marker mode menu.
activates an alternate sweep function when dual channel display is on. In this mode the
analyzer alternates between the two sets of stimulus values and displays the measurement data
of both channels .
6-20 Application and Operation Concepts

The following softkeys are located within the sweep type menu. Among them are the ve
sweep types available .
LIN FREQ (linear frequency sweep).
LOG FREQ (logarithmic frequency sweep).
LIST FREQ (list frequency sweep) provides access to the single/all segment menu.
POWER SWEEP .
CW TIME (CW time sweep).
EDIT LIST allows list frequencies to be entered or modi ed using the edit list menu and edit
subsweep menu.
The following sweep types will function with the interpolated error-correction feature
(described later):
The following sweep types will not function with the interpolated error correction feature
(described later):
Linear Frequency Sweep (Hz)
LIN FREQ softkey activates a linear frequency sweep that is displayed on a standard
graticule with ten equal horizontal divisions . This is the preset default sweep type .
F or a linear sweep , sweep time is combined with the channel's frequency span to compute a
source sweep rate:
sweep rate = (frequency span) / (sweep time)
Since the sweep time may be a ected by various factors , the equation provided here is merely
an indication of the ideal (fastest) sweep rate . If the user-speci ed sweep time is greater
than 15 ms times the number of points , the sweep changes from a continuous ramp sweep
to a stepped CW sweep . Also , for 10 Hz or 30 Hz IF bandwidths the sweep is automatically
converted to a stepped CW sweep .
In the linear frequency sweep mode it is possible , with Option 010, to transform the data for
time domain measurements using the inverse F ourier transform technique .
Application and Operation Concepts 6-21

Logarithmic Frequency Sweep (Hz)
LOG FREQ softkey activates a logarithmic frequency sweep mode . The source is stepped
in logarithmic increments and the data is displayed on a logarithmic graticule . This is slower
than a continuous sweep with the same number of points , and the entered sweep time may
therefore be changed automatically . F or frequency spans of less than two octaves , the sweep
type automatically reverts to linear sweep .
List Frequency Sweep (Hz)
LIST FREQ softkey provides a user-de nable arbitrary frequency list mode . This list is
de ned and modi ed using the edit list menu and the edit subsweep menu. Up to 30 frequency
subsweeps (called \segments") of several di erent types can be speci ed, for a maximum total
of 1632 points . One list is common to both channels . Once a frequency list has been de ned
and a measurement calibration performed on the full frequency list, one or all of the frequency
segments can be measured and displayed without loss of calibration.
LIST FREQ key is pressed, the network analyzer sorts all the de ned frequency
segments into CW points in order of increasing frequency . It then measures each point and
displays a single trace that is a composite of all data taken. If duplicate frequencies exist, the
analyzer makes multiple measurements on identical points to maintain the speci ed number of
points for each subsweep . Since the frequency points may not be distributed evenly across
the display , the display resolution may be uneven, and more compressed in some parts of the
trace than in others . However , the stimulus and response readings of the markers are always
accurate . Because the list frequency sweep is a stepped CW sweep , the sweep time is slower
than for a continuous sweep with the same number of points .
LIST FREQ softkey provides access to the segment menu, which allows you to select
any single segment (
SINGLE SEG SWEEP ) in the frequency list or all of the segments
ALL SEGS SWEEP ) in the frequency list.
Refer to \Modifying List Frequencies" in this section for information on how to enter or modify
the list frequencies . If no list has been entered, the message CAUTION: LIST TABLE EMPTY is
displayed.
A tabular printout of the frequency list data can be obtained using the
in the copy menu.
6-22 Application and Operation Concepts

P ower Sweep (dBm)
POWER SWEEP softkey turns on a power sweep mode that is used to characterize
power-sensitive circuits . In this mode , power is swept at a single frequency , from a start power
value to a stop power value , selected using the
keys and the entry block.
STOP
This feature is convenient for such measurements as gain compression or A GC (automatic gain
NNNNNNNNNNNNNNNNNNNNNN
control) slope . T o set the frequency of the power sweep , use
CW FREQ in the stimulus menu.
The span of the swept power is limited to being equal to or within one of the eight pre-de ned
power ranges . The attenuator will not switch to a di erent power range while in the power
sweep mode . Therefore , when performing a power sweep , power range selection will
automatically switch to the manual mode .
In power sweep , the entered sweep time may be automatically changed if it is less than the
minimum required for the current con guration (number of points , IF bandwidth, averaging,
etc .).
CW Time Sweep (Seconds)
CW TIME softkey turns on a sweep mode similar to an oscilloscope . The analyzer is set to a
single frequency , and the data is displayed versus time . The frequency of the CW time sweep is
CW FREQ in the stimulus menu. In this sweep mode , the data is continuously sampled
at precise , uniform time intervals determined by the sweep time and the number of points
minus 1. The entered sweep time may be automatically changed if it is less than the minimum
required for the current instrument con guration.
In time domain using Option 010, the CW time mode data is translated to frequency domain,
and the x-axis becomes frequency . In this mode , the instrument can be used as a spectrum
analyzer to measure signal purity , or for low frequency (
modulation signals .
In addition to the previous sweep types , there are also two di erent sweep modes . These
can be accessed through the correction menu by pressing
MORE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ALTERNATE A and B
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CHOP A and B . Refer to \Alternate and Chop Sweep Modes" in the \Measurement
Calibration" section.
List frequencies can be entered or modi ed using the edit list and edit subsweep menus .
Application of the functions in these menus is described below .
EDIT LIST softkey within the sweep type menu provides access to the edit list menu.
This menu is used to edit the list of frequency segments (subsweeps) de ned with the edit
subsweep menu, described next. Up to 30 frequency subsweeps can be speci ed, for a
maximum of 1632 points . The segments do not have to be entered in any particular order:
the analyzer automatically sorts them and shows them on the display in increasing order of
start frequency . This menu determines which entry on the list is to be modi ed, while the edit
subsweep menu is used to make changes in the frequency or number of points of the selected
entry .
Application and Operation Concepts 6-23

NNNNNNNNNNN
ADD softkey within the edit list menu will display the edit subsweep menu.
This menu lets you select measurement frequencies arbitrarily . Using this menu it is possible to
de ne the exact frequencies to be measured on a point-by-point basis . F or example the sweep
could include 100 points in a narrow passband, 100 points across a broad stop band, and 50
points across the third harmonic response . The total sweep is de ned with a list of subsweeps .
The frequency subsweeps , or segments , can be de ned in any of the following terms:
The subsweeps can overlap , and do not have to be entered in any particular order . The
analyzer sorts the segments automatically and lists them on the display in order of increasing
start frequency , even if they are entered in center/span format. If duplicate frequencies
exist, the analyzer makes multiple measurements on identical points to maintain the speci ed
number of points for each subsweep . The data is shown on the display as a single trace that is
a composite of all data taken. The trace may appear uneven because of the distribution of the
data points , but the frequency scale is linear across the total range .
Once the list frequencies have been de ned or modi ed, the list frequency sweep mode can be
LIST FREQ softkey in the sweep type menu (see \List Frequency Sweep").
The frequency list parameters can also be saved with an instrument state .
6-24 Application and Operation Concepts


Figure 6-7. Response Function Block
The following response function block keys are used to de ne and control the following
functions of the active channel.
DISPLA Y
MARKER F CTN
The current values for the major response functions of the active channel are displayed in
speci c locations along the top of the display . In addition, certain functions accessed through
the keys in this block are annotated in the status notations area at the left side of the display .
An illustration of the analyzer's display showing the locations of these information labels is
provided in Chapter 1, \HP 8753D Description and Options ."
Application and Operation Concepts 6-25

key provides access to the S-parameter menu which contains softkeys that can be
used to select the parameters or inputs that de ne the type of measurement being performed.
S-parameters (scattering parameters) are a convention used to characterize the way a device
modi es signal ow . A brief explanation of the S-parameters of a two-port device is provided
here . F or additional details refer to Hewlett-P ackard Application Notes A/N 95-1 and A/N 154.
S-parameters are always a ratio of two complex (magnitude and phase) quantities . S-parameter
where the rst number (out) refers to the test-device port where the signal is emerging and
the second number (in) is the test-device port where the signal is incident. F or example , the
identi es the measurement as the complex ratio of the signal emerging at the
test device's port 2 to the signal incident at the test device's port 1.
Figure 6-8 is a representation of the S-parameters of a two-port device , together with an
equivalent owgraph. In the illustration, \a" represents the signal entering the device and \b"
represents the signal emerging. Note that a and b are not related to the A and B input ports on
the analyzer .

Figure 6-8. S-P arameters of a Two-P ort Device
6-26 Application and Operation Concepts

S-parameters are exactly equivalent to the more common description terms below , requiring
only that the measurements be taken with all test device ports properly terminated.
S-P arameter De nition T est Set
= 0 Input re ection coecient FWD
= 0 F orward gain FWD
= 0 Reverse gain REV
= 0 Output re ection coecient REV
The S-P arameter Menu
The S-parameter menu allows you to de ne the input ports and test set direction for
S-parameter measurements . The analyzer automatically switches the direction of the
measurement according to the selections you made in this menu. Therefore , the analyzer can
measure all four S-parameters with a single connection. The S-parameter being measured is
labeled at the top left corner of the display .
The S-parameter menu contains the following softkeys:
ANALOG IN Aux Input
CONVERSION [ ] provides access to the conversion menu.
INPUT PORTS provides access to the input ports menu.
This menu allows you to monitor voltage and frequency nodes , using the analog bus and
internal counter . F or more information, refer to Chapter 10, \Service Key Menus and Error
Messages" in the HP 8753D Network Analyzer Service Guide .
This menu converts the measured re ection or transmission data to the equivalent complex
impedance (Z) or admittance (Y) values . This is not the same as a two-port Y or Z parameter
conversion, as only the measured parameter is used in the equations . Two simple one-port
conversions are available , depending on the measurement con guration.
trace measured as re ection can be converted to equivalent parallel impedance or
admittance using the model and equations shown in Figure 6-9.
Application and Operation Concepts 6-27


Figure 6-9. Re ection Impedance and A dmittance Conversions
In a transmission measurement, the data can be converted to its equivalent series impedance or
admittance using the model and equations shown in Figure 6-10.

Figure 6-10. Transmission Impedance and A dmittance Conversions
A void the use of Smith chart, SWR, and delay formats for display of Z and Y
conversions , as these formats are not easily interpreted.
Input P orts Menu
This menu allows you to de ne the input ports for power ratio measurements , or a single input
for magnitude only measurements of absolute power . Y ou cannot use single inputs for phase or
group delay measurements , or any measurements with averaging activated.
6-28 Application and Operation Concepts

The F ormat Menu
key provides access to the format menu. This menu allows you to select the
appropriate display format for the measured data. The following list identi es which formats
are available by means of which softkeys:
NNNNNNNNNNNNNNNNNNNNNN
LOG MAG
NNNNNNNNNNNNNNNNN
PHASE
NNNNNNNNNNNNNNNNN
DELAY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SMITH CHART
NNNNNNNNNNNNNNNNN
POLAR
NNNNNNNNNNNNNNNNNNNNNNN
LIN MAG
NNNNNNNNNNN
SWR
NNNNNNNNNNNNNN
REAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
IMAGINARY
The analyzer automatically changes the units of measurement to correspond with the displayed
format. Special marker menus are available for the polar and Smith formats , each providing
several di erent marker types for readout of values .
The selected display format of a particular S-parameter or input is assigned to that parameter .
Thus if di erent S-parameters are measured, even if only one channel is used, each parameter
is shown in its selected format each time it is displayed.
The illustrations below show a re ection measurement of a bandpass lter displayed in each of
the available formats .
Log Magnitude F ormat
LOG MAG softkey displays the log magnitude format. This is the standard Cartesian format
used to display magnitude-only measurements of insertion loss , return loss , or absolute power
in dB versus frequency . Figure 6-11 illustrates the bandpass lter re ection data in a log
magnitude format.
Application and Operation Concepts 6-29
The · The


Figure 6-11. Log Magnitude F ormat
PHASE softkey displays a Cartesian format of the phase portion of the data, measured
in degrees . This format displays the phase shift versus frequency . Figure 6-12 illustrates the
phase response of the same lter in a phase-only format.

Figure 6-12. Phase F ormat
Group Delay F ormat
DELAY softkey selects the group delay format, with marker values given in seconds .
Figure 6-13 shows the bandpass lter response formatted as group delay . Group delay
principles are described in the next few pages .
6-30 Application and Operation Concepts
Phase F ormat · The · The


Figure 6-13. Group Delay F ormat
Smith Chart F ormat
SMITH CHART softkey displays a Smith chart format (see Figure 6-14). This is used
in re ection measurements to provide a readout of the data in terms of impedance . The
intersecting dotted lines on the Smith chart represent constant resistance and constant
reactance values , normalized to the characteristic impedance , Z
, of the system. Reactance
values in the upper half of the Smith chart circle are positive (inductive) reactance , and
those in the lower half of the circle are negative (capacitive) reactance . The default marker
readout is in ohms ( ) to measure resistance and reactance (R+jX). A dditional marker types are
available in the Smith marker menu.
The Smith chart is most easily understood with a full scale value of 1.0. If the scale per
division is less than 0.2, the format switches automatically to polar .
If the characteristic impedance of the system is not 50 ohms , modify the impedance value
recognized by the analyzer by pressing
NNNNNNNNNNNNNNNNNNNN
SET Z0 (the impedance value)
.
MORE
An inverted Smith chart format for admittance measurements (Figure 6-14) is also available .
A ccess this by selecting
SMITH CHART in the format menu, and pressing
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNN
MKR MODE MENU
SMITH MKR MENU
G+jB MKR . The Smith chart is inverted and marker values
are read out in siemens (S) to measure conductance and susceptance (G+jB).
Application and Operation Concepts 6-31
The · ·


Figure 6-14. Standard and Inverse Smith Chart F ormats
P olar F ormat
POLAR softkey displays a polar format (see Figure 6-15). Each point on the polar format
corresponds to a particular value of both magnitude and phase . Quantities are read vectorally:
the magnitude at any point is determined by its displacement from the center (which has zero
value), and the phase by the angle counterclockwise from the positive x-axis . Magnitude is
scaled in a linear fashion, with the value of the outer circle usually set to a ratio value of 1.
Since there is no frequency axis , frequency information is read from the markers .
The default marker readout for the polar format is in linear magnitude and phase . A log
magnitude marker and a real/imaginary marker are available in the polar marker menu.

Figure 6-15. P olar F ormat
6-32 Application and Operation Concepts
The

Linear Magnitude F ormat
LIN MAG softkey displays the linear magnitude format (see Figure 6-16). This is a
Cartesian format used for unitless measurements such as re ection coecient magnitude or
transmission coecient magnitude , and for linear measurement units . It is used for display of
conversion parameters and time domain transform data.

Figure 6-16. Linear Magnitude F ormat
SWR softkey reformats a re ection measurement into its equivalent SWR (standing wave
ratio) value (see Figure 6-17 ). SWR is equivalent to (1 + )/(1 ), where is the re ection
coecient. Note that the results are valid only for re ection measurements . If the SWR format
is used for measurements of S
the results are not valid.

Figure 6-17. Typical SWR Display
Application and Operation Concepts 6-33

REAL softkey displays only the real (resistive) portion of the measured data on a Cartesian
format (see Figure 6-18). This is similar to the linear magnitude format, but can show both
positive and negative values . It is primarily used for analyzing responses in the time domain,
and also to display an auxiliary input voltage signal for service purposes .

Figure 6-18. Real F ormat
IMAGINARY softkey displays only the imaginary (reactive) portion of the measured data
on a Cartesian format. This format is similar to the real format except that reactance data is
displayed on the trace instead of impedance data.
6-34 Application and Operation Concepts

F or many networks , the amount of insertion phase is not as important as the linearity of the
phase shift over a range of frequencies . The analyzer can measure this linearity and express
it in two di erent ways: directly , as deviation from linear phase , or as group delay , a derived
value .
Group delay is the measurement of signal transmission time through a test device . It is de ned
as the derivative of the phase characteristic with respect to frequency . Since the derivative
is basically the instantaneous slope (or rate of change of phase with respect to frequency), a
perfectly linear phase shift results in a constant slope , and therefore a constant group delay
(see Figure 6-19).

Figure 6-19. Constant Group Delay
order (deviations from linear) components . The linear component can be attributed to the
electrical length of the test device , and represents the average signal transit time . The higher
order components are interpreted as variations in transit time for di erent frequencies , and
represent a source of signal distortion (see Figure 6-20).

Figure 6-20. Higher Order Phase Shift
The analyzer computes group delay from the phase slope . Phase data is used to nd the phase
change , , over a speci ed frequency aperture , f, to obtain an approximation for the rate of
change of phase with frequency (see Figure 6-21). This value ,
, represents the group delay in
seconds assuming linear phase change over f. It is important that be 180
, or errors will
Application and Operation Concepts 6-35
Group Delay Principles · g ·

result in the group delay data. These errors can be signi cant for long delay devices . Y ou can
verify that is 180
by increasing the number of points or narrowing the frequency span (or
both) until the group delay data no longer changes .

Figure 6-21. Rate of Phase Change V ersus Frequency
When deviations from linear phase are present, changing the frequency step can result in
di erent values for group delay . Note that in this case the computed slope varies as the
aperture f is increased (see Figure 6-22 ). A wider aperture results in loss of the ne grain
variations in group delay . This loss of detail is the reason that in any comparison of group delay
data, it is important to know the aperture that was used to make the measurement.

Figure 6-22. V ariations in Frequency Aperture
In determining the group delay aperture , there is a tradeo between resolution of ne detail
and the e ects of noise . Noise can be reduced by increasing the aperture , but this will tend to
smooth out the ne detail. More detail will become visible as the aperture is decreased, but the
noise will also increase , possibly to the point of obscuring the detail. A good practice is to use a
smaller aperture to assure that small variations are not missed, then increase the aperture to
smooth the trace .
6-36 Application and Operation Concepts

The default group delay aperture is the frequency span divided by the number of points across
the display . T o set the aperture to a di erent value , turn on smoothing in the average menu,
and vary the smoothing aperture . The aperture can be varied up to 20% of the span swept.
Group delay measurements can be made on linear frequency , log frequency , or list frequency
sweep types (not in CW or power sweep). Group delay aperture varies depending on the
frequency spacing and point density , therefore the aperture is not constant in log and list
frequency sweep modes . In list frequency mode , extra frequency points can be de ned to
ensure the desired aperture .
T o obtain a readout of aperture values at di erent points on the trace , turn on a marker . Then
A VG
SMOOTHING APERTURE . Smoothing aperture becomes the active function, and as the
aperture is varied its value in Hz is displayed below the active entry area.
Application and Operation Concepts 6-37
press ·

key provides access to the scale reference menu. Softkeys within this menu
can be used to de ne the scale in which measured data is to be displayed, as well as simulate
phase o set and electrical delay . The following softkeys are located within the scale reference
menu.
AUTO SCALE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SCALE/DIV
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
REFERENCE POSITION
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
REFERENCE VALUE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER ! REFERENCE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ELECTRICAL DELAY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PHASE OFFSET
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
COAXIAL DELAY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
WAVEGUIDE DELAY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ELECTRICAL DELAY softkey adjusts the electrical delay to balance the phase of the test
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
device . This softkey must be used in conjunction with
(with cut-o frequency) in order to identify which type of transmission line the delay is being
added to .
Electrical delay simulates a variable length lossless transmission line , which can be added to or
removed from a receiver input to compensate for interconnecting cables , etc . This function
is similar to the mechanical or analog \line stretchers" of other network analyzers . Delay is
annotated in units of time with secondary labeling in distance for the current velocity factor .
With this feature , and with
MARKER ! DELAY (see \Using Markers"), an equivalent length of
air- lled, lossless transmission line is added or subtracted according to the following formula:
Leng th (meter s) =
(F r eq (M H z ) 1: 20083)
Once the linear portion of the test device's phase has been removed, the equivalent length of
the lossless , transmission line can be read out in the active marker area. If the average relative
) of the test device is known over the frequency span, the length calculation can
be adjusted to indicate the actual length of the test device more closely . This can be done by
entering the relative velocity factor for the test device using the calibrate more menu. The
relative velocity factor for a given dielectric can be calculated by:
V el ocity F actor =
assuming a relative permeability of 1.
6-38 Application and Operation Concepts

key provides access to the display menu, which controls the memory math
functions and leads to other menus associated with display functions .
The following softkeys are located within the display menu:
DUAL CHAN on OFF
DISPLAY: DATA
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DISPLAY: MEMORY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DISPLAY: DATA/MEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DISPLAY: DATA - MEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DISPLAY: DATA ! MEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SPLIT DISP ON off
BEEP DONE ON off
BEEP WARN on OFF
ADJUST DISPLAY
NNNNNNNNNNNNNNNNN
TITLE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
D2/D1 TO D2 on OFF
FREQUENCY BLANK
The analyzer has two available memory traces , one per channel. Memory traces are totally
channel dependent: channel 1 cannot access the channel 2 memory trace or vice versa.
Memory traces can be saved with instrument states: one memory trace can be saved per
channel per saved instrument state . There are up to 31 save/recall registers available , so the
total number of memory traces that can be present is 64 including the two active for the
current instrument state . The memory data is stored as full precision, complex data.
Y ou may not be able to store 31 instrument states if they include a large
amount of calibration data. The calibration data contributes considerably to the
size of the instrument state le and therefore the available memory may be full
prior to lling all 31 registers .
Application and Operation Concepts 6-39
Display Menu · The

DUAL CHAN on OFF softkey toggles between display of both measurement channels or
the active channel only . This is used in conjunction with
SPLIT DISP ON off in the display
more menu to display both channels . With
SPLIT DISP on OFF the two traces are overlaid
on a single graticule (see Figure 6-23a); with
SPLIT DISP ON off the measurement data
is displayed on two half-screen graticules one above the other (see Figure 6-23b). Current
parameters for the two displays are annotated separately .
The stimulus functions of the two channels can also be controlled independently using
COUPLED CH ON off in the stimulus menu. In addition, the markers can be controlled
independently for each channel using
MARKERS: UNCOUPLED in the marker mode menu.
Dual Channel Mode with Decoupled Channel P ower
By decoupling the channel power or port power and using the dual channel mode , you can
simultaneously view two measurements having di erent power levels . However , there are two
con gurations that may not appear to function \properly".
one attenuator is used for both testports , this would cause the attenuator to continuously
switch power ranges .
2 is driving the other test port. This would cause the test port transfer switch to continually
cycle . The instrument will not allow the transfer switch or attenuator to continuously
switch ranges in order to update these measurements without the direct intervention of the
operator .
If one of the above conditions exists , the test set hold mode will engage , and the status
function in only one of the two measurement paths . T o update both measurement setups , press
MENU
MEASURE RESTART . Refer to \Source Attenuator Switch Protection" earlier in this
chapter for more information on this condition.

Figure 6-23. Dual Channel Displays
6-40 Application and Operation Concepts

Two trace math operations are implemented:
NNNNNNNNNNNNNNNNNNNNNNNNNNN
DATA MEM (datamemory)
NNNNNNNNNNNNNNNNNNNNNNNNNNN
(Note that normalization is
DATA MEM .) Memory traces are saved and recalled
and trace math is done immediately after error-correction. This means that any data processing
done after error-correction, including parameter conversion, time domain transformation
(Option 010), scaling, etc ., can be performed on the memory trace . Y ou can also use trace math
as a simple means of error-correction, although that is not its main purpose .
All data processing operations that occur after trace math, except smoothing and gating, are
identical for the data trace and the memory trace . If smoothing or gating is on when a memory
trace is saved, this state is maintained regardless of the data trace smoothing or gating status .
If a memory trace is saved with gating or smoothing on, these features can be turned on or o
in the memory-only display mode .
The actual memory for storing a memory trace is allocated only as needed. The memory trace
is cleared on instrument preset, power on, or instrument state recall.
If sweep mode or sweep range is di erent between the data and memory traces , trace math is
allowed, and no warning message is displayed. If the number of points in the two traces is
di erent, the memory trace is not displayed nor rescaled. However , if the number of points for
the data trace is changed back to the number of points in the memory , the memory trace can
then be displayed.
If trace math or display memory is requested and no memory trace exists , the message
A djusting the Colors of the Display
ADJUST DISPLAY softkey provides access to the adjust display menu. The following
softkeys are located within this menu:
INTENSITY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
BACKGROUND INTENSITY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MODIFY COLORS
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DEFAULT COLORS
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
BLANK DISPLAY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SAVE COLORS
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RECALL COLORS
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
T o adjust the intensity of the display , press
INTENSITY and rotate the front panel knob , use
keys , or use the numerical keypad to set the intensity value between 50 and 100
percent. Lowering the intensity may prolong the life of the LCD .
Application and Operation Concepts 6-41

T o set all the display elements to the factory-de ned default colors , press
DEFAULT COLORS .
PRESET does not reset or change colors to the default color values . However ,
cycling power to the instrument will reset the colors to the default color
values .
BLANK DISPLAY switches o the analyzer display while leaving the instrument in
its current measurement state . This feature may be helpful in prolonging the life of the LCD
in applications where the analyzer is left unattended (such as in an automated test system).
Turning the front panel knob or pressing any front panel key will restore normal display
operation.
Saving Modi ed Colors
T o save a modi ed color set, press
SAVE COLORS . Modi ed colors are not part of a saved
instrument state and are lost unless saved using these softkeys .
Recalling Modi ed Colors
T o recall the previously saved color set, press
RECALL COLORS .
The Modify Colors Menu
MODIFY COLORS softkey within the adjust display menu provides access to the modify
colors menu.
The modify colors menu allows you to adjust the colors on your analyzer's display . The default
colors in this instrument have been scienti cally chosen to maximize your ability to discern the
di erence between the colors , and to comfortably and e ectively view the colors . These colors
are recommended for normal use because they will provide a suitable contrast that is easy to
view for long periods of time .
Y ou may choose to change the default colors to suit environmental needs , individual
preferences , or to accommodate color de cient vision. Y ou can use any of the available colors
for any of the seven display elements listed by the softkey names below:
CH1 DATA/LIMIT LN
NNNNNNNNNNNNNNNNNNNNNNN
CH1 MEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CH2 DATA/LIMIT LN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CH2 MEM/REF LINE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
GRATICULE/TEXT
NNNNNNNNNNNNNNNNNNNNNNN
WARNING
NNNNNNNNNNNNNN
TEXT
T o change the color of a display elements , press the softkey for that element (such as
CH1 DATA ). Then press
TINT and turn the analyzer front panel knob , use the step keys or the
numeric keypad, until the desired color appears .
6-42 Application and Operation Concepts

Maximum viewing angle with the LCD display is achieved when primary colors
or a combination of them are selected at full brightness (100%). The following
table lists the recommended colors and their corresponding tint numbers .
T able 6-2. Display Colors with Maximum Viewing Angle
Display Color Tint Brightness Color
Red 0 100 100
Y ellow 17 100 100
Green 33 100 100
Cyan 50 100 100
Blue 67 100 100
Magenta 83 100 100
Color is comprised of three parameters:
Tint: The continuum of hues on the color wheel, ranging from red, through green and blue ,
and back to red.
Brightness: A measure of the brightness of the color .
Color: The degree of whiteness of the color . A scale from white to pure color .
The most frequently occurring color de ciency is the inability to distinguish red, yellow ,
and green from one another . Confusion between these colors can usually be eliminated by
increasing the brightness between the colors . T o accomplish this , press the
softkey and turn the analyzer front panel knob . If additional adjustment is needed, vary the
degree of whiteness of the color . T o accomplish this , press the
COLOR softkey and turn the
analyzer front panel knob .
Color changes and adjustments remain in e ect until changed again in these
menus or the analyzer is powered o and then on again. Cycling the power
changes all color adjustments to default values . Preset does not a ect color
selection.
Application and Operation Concepts 6-43
White 100 0

key is used to access three di erent noise reduction techniques: sweep-to-sweep
averaging, display smoothing, and variable IF bandwidth. All of these can be used
simultaneously . A veraging and smoothing can be set independently for each channel, and the
IF bandwidth can be set independently if the stimulus is uncoupled.
The following softkeys are located within the averaging menu:
AVERAGING RESTART
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AVERAGING FACTOR
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SMOOTHING APERTURE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
IF BW [ ]
A veraging computes each data point based on an exponential average of consecutive sweeps
weighted by a user-speci ed averaging factor . Each new sweep is averaged into the trace until
the total number of sweeps is equal to the averaging factor , for a fully averaged trace . Each
point on the trace is the vector sum of the current trace data and the data from the previous
sweep . A high averaging factor gives the best signal-to-noise ratio , but slows the trace update
time . Doubling the averaging factor reduces the noise by 3 dB . A veraging is used for ratioed
measurements: if it is attempted for a single-input measurement (e .g. A or B), the message
e ect of averaging on a log magnitude format trace .
If you switch power ranges with averaging on, the average will restart.

Figure 6-24. E ect of A veraging on a Trace
6-44 Application and Operation Concepts

Smoothing (similar to video ltering) averages the formatted active channel data over a portion
of the displayed trace . Smoothing computes each displayed data point based on one sweep
only , using a moving average of several adjacent data points for the current sweep . The
smoothing aperture is a percent of the swept stimulus span, up to a maximum of 20%.
Rather than lowering the noise oor , smoothing nds the mid-value of the data. Use it
to reduce relatively small peak-to-peak noise values on broadband measured data. Use a
suciently high number of display points to avoid misleading results . Do not use smoothing for
measurements of high resonance devices or other devices with wide trace variations , as it will
introduce errors into the measurement.
Smoothing is used with Cartesian and polar display formats . It is also the primary way to
control the group delay aperture , given a xed frequency span. (Refer to \Group Delay
Principles" earlier in this section.) In polar display format, large phase shifts over the
smoothing aperture will cause shifts in amplitude , since a vector average is being computed.
Figure 6-25 illustrates the e ect of smoothing on a log magnitude format trace .

Figure 6-25. E ect of Smoothing on a Trace
IF bandwidth reduction lowers the noise oor by digitally reducing the receiver input
bandwidth. It works in all ratio and non-ratio modes . It has an advantage over averaging as it
reliably lters out unwanted responses such as spurs , odd harmonics , higher frequency spectral
noise , and line-related noise . Sweep-to-sweep averaging, however , is better at ltering out very
low frequency noise . A tenfold reduction in IF bandwidth lowers the measurement noise oor
by about 10 dB . Bandwidths less than 300 Hz provide better harmonic rejection than higher
bandwidths .
Another di erence between sweep-to-sweep averaging and variable IF bandwidth is the sweep
time . A veraging displays the rst complete trace faster but takes several sweeps to reach a
fully averaged trace . IF bandwidth reduction lowers the noise oor in one sweep , but the
sweep time may be slower . Figure 6-26 illustrates the di erence in noise oor between a trace
measured with a 3000 Hz IF bandwidth and with a 10 Hz IF bandwidth.
Application and Operation Concepts 6-45
Smoothing · IF Bandwidth Reduction


Figure 6-26. IF Bandwidth Reduction
Another capability that can be used for e ective noise reduction is the marker
statistics function, which computes the average value of part or all of the
formatted trace .
If your instrument is equipped with Option 085 (High P ower System), another
way of increasing dynamic range is to increase the input power to the test
device using a booster ampli er .
6-46 Application and Operation Concepts

key displays a movable active marker on the screen and provides access to a
series of menus to control up to ve display markers for each channel. Markers are used to
obtain numerical readings of measured values . They also provide capabilities for reducing
measurement time by changing stimulus parameters , searching the trace for speci c values , or
statistically analyzing part or all of the trace . Figure 6-27 illustrates the displayed trace with all
markers on and marker 2 the active marker .

Figure 6-27. Markers on Trace
Markers have a stimulus value (the x-axis value in a Cartesian format) and a response value
(the y-axis value in a Cartesian format). In a polar or Smith chart format, the second part of a
complex data pair is also provided as an auxiliary response value . When a marker is activated
and no other function is active , its stimulus value is displayed in the active entry area and can
be controlled with the knob , the step keys , or the numeric keypad. The active marker can be
moved to any point on the trace , and its response and stimulus values are displayed at the top
right corner of the graticule for each displayed channel, in units appropriate to the display
format. The displayed marker response values are valid even when the measured data is above
or below the range displayed on the graticule .
points . Or , they can be set to read only discrete measured points . The markers for the two
channels normally have the same stimulus values , or they can be uncoupled so that each
channel has independent markers , regardless of whether stimulus values are coupled or dual
channel display is on.
If both data and memory are displayed, the marker values apply to the data trace . If only
memory is displayed, the marker values apply to the memory trace . In a memory math display
(data/memory or datamemory), the marker values apply to the trace resulting from the
memory math function.
Application and Operation Concepts 6-47
Markers · The

With the use of a reference marker , a delta marker mode is available that displays both the
stimulus and response values of the active marker relative to the reference . Any of the ve
markers or a xed point can be designated as the delta reference marker . If the delta reference
is one of the ve markers , its stimulus value can be controlled by the user and its response
value is the value of the trace at that stimulus value . If the delta reference is a xed marker ,
both its stimulus value and its response value can be set arbitrarily anywhere in the display
area (not necessarily on the trace).
Markers can be used to search for the trace maximum or minimum point or any other point
on the trace . The ve markers can be used together to search for speci ed bandwidth cuto
points and calculate the bandwidth and Q values . Statistical analysis uses markers to provide a
readout of the mean, standard deviation, and peak-to-peak values of all or part of the trace .
Basic marker operations are available in the menus accessed from the
key . The
marker search and statistical functions , together with the capability for quickly changing
stimulus parameters with markers , are provided in the menus accessed from the
key .
key provides access to the marker menu. This menu allows you to turn the
display markers on or o , to designate the active marker , and to gain access to the delta
marker menu and the xed marker menu.
MODE MENU softkey within the marker menu provides access to the delta mode menu.
The delta reference is shown on the display as a small triangle , smaller than the inactive
marker triangles . If one of the markers is the reference , the triangle appears next to the
marker number on the trace .
The marker values displayed in this mode are the stimulus and response values of the active
marker minus the reference marker . If the active marker is also designated as the reference
marker , the marker values are zero .
Fixed Marker Menu. The
FIXED MKR POSTION softkey within the delta mode menu provides
access to the xed marker menu. This menu is used to set the position of a xed reference
marker , indicated on the display by a small triangle . Both the stimulus value and the
response value of the xed marker can be set arbitrarily anywhere in the display area, and
need not be on the trace . The units are determined by the display format, the sweep type , and
the marker type .
There are two ways to turn on the xed marker . One way is with the
softkey in the delta marker menu. The other is with the
MKR ZERO function in the marker
menu, which puts a xed reference marker at the present active marker position and makes
the marker stimulus and response values at that position equal to zero .
The softkeys in this menu make the values of the xed marker the active function. The marker
readings in the top right corner of the graticule are the stimulus and response values of the
active marker minus the xed reference marker . Also displayed in the top right corner is the
The stimulus value , response value , and auxiliary response value (the second part of a complex
data pair) can be individually examined and changed. This allows active marker readings that
are relative in amplitude yet absolute in frequency , or any combination of relative/absolute
readouts . F ollowing a
MKR ZERO operation, this menu can be used to reset any of the xed
marker values to absolute zero for absolute readings of the subsequent active marker values .
6-48 Application and Operation Concepts

If the format is changed while a xed marker is on, the xed marker values become invalid.
F or example , if the value o set is set to 10 dB with a log magnitude format, and the format
is then changed to phase , the value o set becomes 10 degrees . However , in polar and Smith
chart formats , the speci ed values remain consistent between di erent marker types for those
formats . Thus an R+jX marker set on a Smith chart format will retain the equivalent values if
it is changed to any of the other Smith chart markers .
key provides access to the marker function menu. This menu provides
softkeys that use markers to quickly modify certain measurement parameters without going
through the usual key sequence . In addition, it provides access to two additional menus used
for searching the trace and for statistical analysis .
MARKER ! functions change certain stimulus and response parameters to make them
equal to the current active marker value . Use the knob or the numeric keypad to move the
marker to the desired position on the trace , and press the appropriate softkey to set the
speci ed parameter to that trace value . When the values have been changed, the marker can
again be moved within the range of the new parameters .
MKR SEARCH [ ] softkey within the marker function menu provides access to the
marker search menu. This menu is used to search the trace for a speci c amplitude-related
point, and place the marker on that point. The capability of searching for a speci ed
bandwidth is also provided. Tracking is available for a continuous sweep-to-sweep search. If
there is no occurrence of a speci ed value or bandwidth, the message TARGET VALUE NOT FOUND
is displayed.
T arget Menu. The
TARGET softkey within the marker search menu provides access to the
target menu. This menu lets you place the marker at a speci ed target response value on
the trace , and provides search right and search left options . If there is no occurrence of the
speci ed value , the message TARGET VALUE NOT FOUND is displayed.
MKR MODE MENU softkey within the marker function menu provides access to the marker
mode menu.This menu provides di erent marker modes and leads to the following two menus:
P olar Marker Menu. This menu is used only with a polar display format, selectable using the
key . In a polar format, the magnitude at the center of the circle is zero and the outer
circle is the full scale value set in the scale reference menu. Phase is measured as the angle
at the positive x-axis . The analyzer automatically calculates di erent
mathematical forms of the marker magnitude and phase values , selected using the softkeys in
this menu. Marker frequency is displayed in addition to other values regardless of the selection
of marker type .
Smith Marker Menu. This menu is used only with a Smith chart format, selected from
the format menu. The analyzer automatically calculates di erent mathematical forms of
the marker magnitude and phase values , selected using the softkeys in this menu. Marker
frequency is displayed in addition to other values for all marker types .
Application and Operation Concepts 6-49

Measurement calibration is an accuracy enhancement procedure that e ectively removes the
system errors that cause uncertainty in measuring a test device . It measures known standard
devices , and uses the results of these measurements to characterize the system.
This section discusses the following topics:
de nition of accuracy enhancement
causes of measurement errors
characterization of microwave systematic errors
e ectiveness of accuracy enhancement
correcting for measurement errors
ensuring a valid calibration
calibrating for non-insertable devices
What is A ccuracy Enhancement?
A perfect measurement system would have in nite dynamic range , isolation, and directivity
characteristics , no impedance mismatches in any part of the test setup , and at frequency
response . In any high frequency measurement there are measurement errors associated with
the system that contribute uncertainty to the results . P arts of the measurement setup such as
interconnecting cables and signal-separation devices (as well as the analyzer itself) all introduce
variations in magnitude and phase that can mask the actual performance of the test device .
V ector accuracy enhancement, also known as measurement calibration or error-correction,
provides the means to simulate a nearly perfect measurement system.
F or example , crosstalk due to the channel isolation characteristics of the analyzer can
contribute an error equal to the transmission signal of a high-loss test device . F or re ection
measurements , the primary limitation of dynamic range is the directivity of the test setup .
The measurement system cannot distinguish the true value of the signal re ected by the test
device from the signal arriving at the receiver input due to leakage in the system. F or both
transmission and re ection measurements , impedance mismatches within the test setup cause
measurement uncertainties that appear as ripples superimposed on the measured data.
Error-correction simulates an improved analyzer system. During the measurement calibration
process , the analyzer measures the magnitude and phase responses of known standard devices ,
and compares the measurement with actual device data. The analyzer uses the results to
characterize the system and e ectively remove the system errors from the measurement data
of a test device , using vector math capabilities internal to the network analyzer .
When you use a measurement calibration, the dynamic range and accuracy of the measurement
are limited only by system noise and stability , connector repeatability , and the accuracy to
which the characteristics of the calibration standards are known.
6-50 Application and Operation Concepts

What Causes Measurement Errors?
Network analysis measurement errors can be separated into systematic , random, and drift
errors .
Correctable systematic errors are the repeatable errors that the system can measure . These are
errors due to mismatch and leakage in the test setup , isolation between the reference and test
signal paths , and system frequency response .
The system cannot measure and correct for the non-repeatable random and drift errors . These
errors a ect both re ection and transmission measurements . Random errors are measurement
variations due to noise and connector repeatability . Drift errors include frequency drift,
temperature drift, and other physical changes in the test setup between calibration and
measurement.
The resulting measurement is the vector sum of the test device response plus all error terms .
The precise e ect of each error term depends upon its magnitude and phase relationship to the
actual test device response .
In most high frequency measurements the systematic errors are the most signi cant source of
measurement uncertainty . Since each of these errors can be characterized, their e ects can be
e ectively removed to obtain a corrected value for the test device response . F or the purpose of
vector accuracy enhancement these uncertainties are quanti ed as directivity , source match,
load match, isolation (crosstalk), and frequency response (tracking). Each of these systematic
errors is described below .
Random and drift errors cannot be precisely quanti ed, so they must be treated as producing a
cumulative uncertainty in the measured data.
Normally a device that can separate the reverse from the forward traveling waves (a
directional bridge or coupler) is used to detect the signal re ected from the test device . Ideally
the coupler would completely separate the incident and re ected signals , and only the re ected
signal would appear at the coupled output, as illustrated in Figure 6-28 a.

Figure 6-28. Directivity
However , an actual coupler is not perfect, as illustrated in Figure 6-28b . A small amount of
the incident signal appears at the coupled output due to leakage as well as re ection from the
termination in the coupled arm. Also , re ections from the coupler output connector appear at
the coupled output, adding uncertainty to the signal re ected from the device . The gure of
merit for how well a coupler separates forward and reverse waves is directivity . The greater
the directivity of the device , the better the signal separation. System directivity is the vector
sum of all leakage signals appearing at the analyzer receiver input. The error contributed by
directivity is independent of the characteristics of the test device and it usually produces the
major ambiguity in measurements of low re ection devices .
Application and Operation Concepts 6-51
Directivity

Source match is de ned as the vector sum of signals appearing at the analyzer receiver input
due to the impedance mismatch at the test device looking back into the source , as well as to
adapter and cable mismatches and losses . In a re ection measurement, the source match error
signal is caused by some of the re ected signal from the test device being re ected from the
source back toward the test device and re-re ected from the test device (Figure 6-29 ). In a
transmission measurement, the source match error signal is caused by re ection from the test
device that is re-re ected from the source . Source match is most often given in terms of return
loss in dB: thus the larger the number , the smaller the error .

Figure 6-29. Source Match
The error contributed by source match is dependent on the relationship between the actual
input impedance of the test device and the equivalent match of the source . It is a factor
in both transmission and re ection measurements . Source match is a particular problem in
measurements where there is a large impedance mismatch at the measurement plane . (F or
example , re ection devices such as lters with stop bands .)
Load match error results from an imperfect match at the output of the test device . It is caused
by impedance mismatches between the test device output port and port 2 of the measurement
system. As illustrated in Figure 6-30 , some of the transmitted signal is re ected from port
2 back to the test device . A portion of this wave may be re-re ected to port 2, or part may
be transmitted through the device in the reverse direction to appear at port 1. If the test
device has low insertion loss (for example a lter pass band), the signal re ected from port 2
and re-re ected from the source causes a signi cant error because the test device does not
attenuate the signal signi cantly on each re ection. Load match is usually given in terms of
return loss in dB: thus the larger the number , the smaller the error .

Figure 6-30. Load Match
6-52 Application and Operation Concepts
Source Match · Load Match

The error contributed by load match is dependent on the relationship between the actual
output impedance of the test device and the e ective match of the return port (port 2). It is
a factor in all transmission measurements and in re ection measurements of two-port devices .
The interaction between load match and source match is less signi cant when the test device
insertion loss is greater than about 6 dB . However , source match and load match still interact
with the input and output matches of the DUT , which contributes to transmission measurement
errors . (These errors are largest for devices with highly re ective output ports .)
Leakage of energy between analyzer signal paths contributes to error in a transmission
measurement, much like directivity does in a re ection measurement. Isolation is the vector
sum of signals appearing at the analyzer samplers due to crosstalk between the reference and
test signal paths . This includes signal leakage within the test set and in both the RF and IF
sections of the receiver .
The error contributed by isolation depends on the characteristics of the test device . Isolation
is a factor in high-loss transmission measurements . However , analyzer system isolation is more
than sucient for most measurements , and correction for it may be unnecessary .
F or measuring devices with high dynamic range , accuracy enhancement can provide
improvements in isolation that are limited only by the noise oor . Generally , the isolation falls
below the noise oor , therefore , when performing an isolation calibration you should use a
noise reduction function such as averaging or reduce the IF bandwidth.
This is the vector sum of all test setup variations in which magnitude and phase change as a
function of frequency . This includes variations contributed by signal-separation devices , test
cables , adapters , and variations between the reference and test signal paths . This error is a
factor in both transmission and re ection measurements .
F or further explanation of systematic error terms and the way they are combined and
represented graphically in error models , refer to the \Characterizing Microwave Systematic
Errors" next.
Characterizing Microwave Systematic Errors
One-P ort Error Model
In a measurement of the re ection coecient (magnitude and phase) of a test device , the
measured data di ers from the actual, no matter how carefully the measurement is made .
Directivity , source match, and re ection signal path frequency response (tracking) are the
major sources of error (see Figure 6-31).
Application and Operation Concepts 6-53
Isolation (Crosstalk) · Frequency Response (Tracking)


Figure 6-31. Sources of Error in a Re ection Measurement
T o characterize the errors , the re ection coecient is measured by rst separating the incident
signal (I) from the re ected signal (R), then taking the ratio of the two values (see Figure 6-32).
Ideally , (R) consists only of the signal re ected by the test device (S
actual).

Figure 6-32. Re ection Coecient
However , all of the incident signal does not always reach the unknown (see Figure 6-33).
Some of (I) may appear at the measurement system input due to leakage through the test
set or through a signal separation device . Also , some of (I) may be re ected by imperfect
adapters between a signal separation device and the measurement plane . The vector sum of
the leakage and the miscellaneous re ections is the e ective directivity , E
. Understandably ,
the measurement is distorted when the directivity signal combines vectorally with the actual
re ected signal from the unknown, S
6-54 Application and Operation Concepts
, for S


Figure 6-33. E ective Directivity E
Since the measurement system test port is never exactly the characteristic impedance
(50 ohms), some of the re ected signal bounces o the test port, or other impedance transitions
further down the line , and back to the unknown, adding to the original incident signal (I). This
e ect causes the magnitude and phase of the incident signal to vary as a function of S
frequency . Leveling the source to produce a constant incident signal (I) reduces this error , but
since the source cannot be exactly leveled at the test device input, leveling cannot eliminate
all power variations . This re-re ection e ect and the resultant incident power variation are
caused by the source match error , E
(see Figure 6-34 ).

Figure 6-34. Source Match E
Frequency response (tracking) error is caused by variations in magnitude and phase atness
versus frequency between the test and reference signal paths . These are due mainly to coupler
roll o , imperfectly matched samplers , and di erences in length and loss between the incident
and test signal paths . The vector sum of these variations is the re ection signal path tracking
(see Figure 6-35).
Application and Operation Concepts 6-55
and · error , E


Figure 6-35. Re ection Tracking E
These three errors are mathematically related to the actual data, S
, and measured data,
S
, by the following equation:
( S
11 A
RF
11 M
= E
E
S
D F
(1 E
11 A
S F
S
If the value of these three \E" errors and the measured test device response were known for
each frequency , the above equation could be solved for S
to obtain the actual test device
response . Because each of these errors changes with frequency , their values must be known
at each test frequency . These values are found by measuring the system at the measurement
plane using three independent standards whose S
is known at all frequencies .
The rst standard applied is a \perfect load," which makes S
= 0 and essentially measures
directivity (see Figure 6-36). \P erfect load" implies a re ectionless termination at the
measurement plane . All incident energy is absorbed. With S
= 0 the equation can be solved
, the directivity term. In practice , of course , the \perfect load" is dicult to achieve ,
although very good broadband loads are available in the HP 8753D compatible calibration kits .

Figure 6-36. \P erfect Load" T ermination
Since the measured value for directivity is the vector sum of the actual directivity plus
the actual re ection coecient of the \perfect load," any re ection from the termination
represents an error . System e ective directivity becomes the actual re ection coecient of the
near \perfect load" (see Figure 6-37 ). In general, any termination having a return loss value
greater than the uncorrected system directivity reduces re ection measurement uncertainty .
6-56 Application and Operation Concepts


Figure 6-37. Measured E ective Directivity
Next, a short circuit termination whose response is known to a very high degree is used to
establish another condition (see Figure 6-38).

Figure 6-38. Short Circuit T ermination
The open circuit gives the third independent condition. In order to accurately model the phase
variation with frequency due to fringing capacitance from the open connector , a specially
designed shielded open circuit is used for this step . (The open circuit capacitance is di erent
with each connector type .) Now the values for E
, directivity , E
, source match, and E
SF
RF
re ection frequency response , are computed and stored (see Figure 6-39).
Application and Operation Concepts 6-57
,


Figure 6-39. Open Circuit T ermination
This completes the calibration procedure .
6-58 Application and Operation Concepts

Now the unknown is measured to obtain a value for the measured response , S
frequency (see Figure 6-40).

Figure 6-40. Measured S
This is the one-port error model equation solved for S
. Since the three errors and S
11M
now known for each test frequency , S
can be computed as follows:
D F
11 A
= S
11 M
E
S
( S
E
) + E
E
S F
11 M
D F
RF
F or re ection measurements on two-port devices , the same technique can be applied, but
the test device output port must be terminated in the system characteristic impedance . This
termination should have as low a re ection coecient as the load used to determine directivity .
The additional re ection error caused by an improper termination at the test device's output
port is not incorporated into the one-port error model.
Two-P ort Error Model
The error model for measurement of the transmission coecients (magnitude and phase) of a
two-port device is derived in a similar manner . The potential sources of error are frequency
response (tracking), source match, load match, and isolation (see Figure 6-41). These errors are
e ectively removed using the full two-port error model.
Application and Operation Concepts 6-59


Figure 6-41. Major Sources of Error
The transmission coecient is measured by taking the ratio of the incident signal (I) and the
transmitted signal (T) (see Figure 6-42 ). Ideally , (I) consists only of power delivered by the
source , and (T) consists only of power emerging at the test device output.

Figure 6-42. Transmission Coecient
As in the re ection model, source match can cause the incident signal to vary as a function
of test device S
. Also , since the test setup transmission return port is never exactly the
characteristic impedance , some of the transmitted signal is re ected from the test set port 2,
and from other mismatches between the test device output and the receiver input, to return
to the test device . A portion of this signal may be re-re ected at port 2, thus a ecting S
or part may be transmitted through the device in the reverse direction to appear at port 1,
thus a ecting S
. This error term, which causes the magnitude and phase of the transmitted
signal to vary as a function of S
, is called load match, E
(see Figure 6-43).
LF
6-60 Application and Operation Concepts
,


Figure 6-43. Load Match E
The measured value , S
, consists of signal components that vary as a function of the
as well as E
, so the input and output re ection
11A
LF
22A
coecients of the test device must be measured and stored for use in the S
computation. Thus , the test setup is calibrated as described above for re ection to establish
the directivity , E
, source match, E
, and re ection frequency response , E
, terms for the
SF
RF
re ection measurements .
Now that a calibrated port is available for re ection measurements , the thru is connected and
, is determined by measuring the re ection coecient of the thru connection.
Transmission signal path frequency response is then measured with the thru connected. The
data is corrected for source and load match e ects , then stored as transmission frequency
It is very important that the exact electrical length of the thru be known.
Most calibration kits assume a zero length thru. F or some connection types
such as Type-N, this implies one male and one female port. If the test system
requires a non-zero length thru, for example , one with two male test ports , the
exact electrical delay of the thru adapter must be used to modify the built-in
calibration kit de nition of the thru.
, represents the part of the incident signal that appears at the receiver without
actually passing through the test device (see Figure 6-44). Isolation is measured with the test
set in the transmission con guration and with terminations installed at the points where the
test device will be connected.
Application and Operation Concepts 6-61


Figure 6-44. Isolation E
Thus there are two sets of error terms , forward and reverse , with each set consisting of six
error terms , as follows:
DR
XF
XR
SF
SR
LF
LR
TF
TR
Re ection Tracking, E
RR
The analyzer's test set can measure both the forward and reverse characteristics of the test
device without you having to manually remove and physically reverse the device . The full
two-port error model illustrated in Figure 6-45 depicts how the analyzer e ectively removes
both the forward and reverse error terms for transmission and re ection measurements .
6-62 Application and Operation Concepts


Figure 6-45. Full Two-P ort Error Model
Figure 6-46 shows the full two-port error model equations for all four S-parameters of a
two-port device . Note that the mathematics for this comprehensive model use all forward and
reverse error terms and measured values . Thus , to perform full error-correction for any one
parameter , all four S-parameters must be measured.
Applications of these error models are provided in the calibration procedures described in
Chapter 5, \Optimizing Measurement Results ."
Application and Operation Concepts 6-63


Figure 6-46. Full Two-P ort Error Model Equations
In addition to the errors removed by accuracy enhancement, other systematic errors exist
due to limitations of dynamic accuracy , test set switch repeatability , and test cable stability .
These , combined with random errors , also contribute to total system measurement uncertainty .
Therefore , after accuracy enhancement procedures are performed, residual measurement
uncertainties remain.
6-64 Application and Operation Concepts

Calibration procedures are parameter-speci c , rather than channel-speci c . When a parameter
is selected, the instrument checks the available calibration data, and uses the data found
for that parameter . F or example , if a transmission response calibration is performed for
B/R, and an S
1-port calibration for A/R, the analyzer retains both calibration sets and
corrects whichever parameter is displayed. Once a calibration has been performed for a
speci c parameter or input, measurements of that parameter remain calibrated in either
channel, as long as stimulus values are coupled. In the response and response and isolation
calibrations , the parameter must be selected before calibration: other correction procedures
select parameters automatically . Changing channels during a calibration procedure invalidates
the part of the procedure already performed.
In calibration procedures that require measurement of several di erent devices , for example
a short, an open, and a load, the order in which the devices are measured is not critical.
NNNNNNNNNNNNN
Any standard can be re-measured, until the
DONE key is pressed. The change in trace during
measurement of a standard is normal.
Response and response and isolation calibrations require measurement of only one standard
device . If more than one device is measured, only the data for the last device is retained.
Isolation calibration can be omitted for most measurements , except where high dynamic range
is a consideration. Use the following guidelines . When the measurement requires a dynamic
range of:
90 dB: Omit isolation calibration for most measurements .
dBm. F or this isolation calibration, averaging should be turned on with an averaging factor
at least four times the measurement averaging factor . F or example , use use an averaging
factor of 16 for the isolation calibration, and then reduce the averaging factor to four for the
measurement after calibration.
100 dB: Same as above , but alternate mode should be used. See page 5-53.
Y ou should save the calibration data, either in the internal non-volatile memory or on a disk.
If you do not save it, it will be lost if you select another calibration procedure for the same
channel, or if you change stimulus values . Instrument preset, power on, and instrument state
recall will also clear the calibration data.
Application and Operation Concepts 6-65

During measurement calibration, the analyzer measures actual, well-de ned standards and
mathematically compares the results with ideal \models" of those standards . The di erences
are separated into error terms which are later removed during error-correction. Most of the
di erences are due to systematic errors - repeatable errors introduced by the analyzer , test set,
and cables - which are correctable .
The standard devices required for system calibration are available in compatible calibration
kits with di erent connector types . Each kit contains at least one short circuit, one open
circuit, and an impedance-matched load. In kits that require adapters for interface to
the test set ports , the adapters are phase-matched for calibration prior to measurement of
non-insertable and non-reversible devices . Other standard devices can be used by specifying
their characteristics in a user-de ned kit, as described later in this section under \Modifying
Calibration Kits ."
The accuracy improvement of the correction is limited by the quality of the standard devices ,
and by the connection techniques used. F or maximum accuracy , use a torque wrench for nal
connections .
Frequency Response of Calibration Standards
In order for the response of a reference standard to show as a dot on the smith chart display
format, it must have no phase shift with respect to frequency . Standards that exhibit such
\perfect" response are the following:
7-mm short (with no o set)
type-N male short (with no o set)
There are two reasons why other types of reference standards show phase shift after
calibration:
The reference plane of the standard is electrically o set from the mating plane of the test
port. Such devices exhibit the properties of a small length of transmission line , including a
certain amount of phase shift.
The standard is an open termination, which by de nition exhibits a certain amount of fringe
capacitance (and therefore phase shift). Open terminations which are o set from the mating
plane will exhibit a phase shift due to the o set in addition to the phase shift caused by the
fringe capacitance .
The most important point to remember is that these properties will not a ect your
measurements . The analyzer compensates for them during measurement. As a result, if these
standards are measured after a calibration, they will not appear to be \perfect" shorts or
opens . This is an indication that your analyzer is working properly and that it has successfully
performed a calibration. Figure 6-47 shows sample displays of various calibration standards
after calibration.
6-66 Application and Operation Concepts
The Calibration Standards

Some standards have reference planes that are electrically o set from the mating plane of the
test port. These devices will show a phase shift with respect to frequency . T able 6-3 shows
which reference devices exhibit an electrical o set phase shift. The amount of phase shift can
be calculated with the formula:
= (360 x f x l)/c where:
l = electrical length of the o set
c = speed of light (3 x 10
All open circuit terminations exhibit a phase shift over frequency due to fringe capacitance .
O set open circuits have increased phase shift because the o set acts as a small length of
transmission line . Refer to T able 6-3.
T able 6-3. Calibration Standard Types and Expected Phase Shift
T est P ort
Standard Type Expected Phase Shift
3.5-mm male O set Short
(360 f l )
3.5-mm female
2.4-mm male
2.4-mm female
3.5-mm male O set Open
(360 f l )
3.5-mm female
2.4-mm male
2.4-mm female
(360 f l )
Application and Operation Concepts 6-67


Figure 6-47. Typical Responses of Calibration Standards after Calibration
6-68 Application and Operation Concepts

How E ective Is A ccuracy Enhancement?
The uncorrected performance of the analyzer is sucient for many measurements . However ,
the vector accuracy enhancement procedures described in Chapter 5, \Optimizing Measurement
Results ," will provide a much higher level of accuracy . Figure 6-48 through Figure 6-50
illustrate the improvements that can be made in measurement accuracy by using a more
complete calibration routine . Figure 6-48a shows a measurement in log magnitude format with
a response calibration only . Figure 6-48b shows the improvement in the same measurement
one-port calibration. Figure 6-49a shows the measurement on a Smith chart with
response calibration only , and Figure 6-49b shows the same measurement with an S
calibration.

Figure 6-48. Response versus S
1-P ort Calibration on Log Magnitude F ormat
Application and Operation Concepts 6-69
using an S · one-port


Figure 6-49. Response versus S
1-P ort Calibration on Smith Chart
Figure 6-50 shows the response of a device in a log magnitude format, using a response
calibration in Figure 6-50a and a full two-port calibration in Figure 6-50b .

Figure 6-50. Response versus Full Two-P ort Calibration
6-70 Application and Operation Concepts

Correcting for Measurement Errors
key provides access to the correction menu which leads to a series of menus that
implement the error-correction concepts described in this section. A ccuracy enhancement
(error-correction) is performed as a calibration step before you measure a test device . When
key is pressed, the correction menu is displayed.
The following softkeys are located within the correction menu:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CALIBRATE MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RESUME CAL SEQUENCE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RECEIVER CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL KIT [ ]
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PWRMTR CAL [OFF]
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PORT EXTENSIONS
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
VELOCITY FACTOR
NNNNNNNNNNNNNNNNNNNN
SET Z
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TEST SET SW [ ]
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ALTERNATE A and B
CHOP A and B
Ensuring a V alid Calibration
Unless interpolated error-correction is on, measurement calibrations are valid only for a
speci c stimulus state , which must be set before a calibration has begun. The stimulus state
consists of the selected frequency range , number of points , sweep time , output power , and
sweep type . Changing the frequency range , number of points , or sweep type with correction
on invalidates the calibration and turns it o . Changing the sweep time or output power
changes the status notation Cor at the left of the screen to C? , to indicate that the calibration
is in question. If correction is turned o or in question after the stimulus changes are made ,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CORRECTION ON off recalls the original stimulus state for the current calibration.
Application and Operation Concepts 6-71

Y ou can activate the interpolated error-correction feature with the
INTERPOL ON off softkey .
This feature allows you to select a subset of the frequency range or a di erent number of
points without recalibration. When interpolation is on, the system errors for the newly selected
frequencies are calculated from the system errors of the original calibration.
System performance is unspeci ed when using interpolated error-correction. The quality
of the interpolated error-correction is dependent on the amount of phase shift and the
amplitude change between measurement points . If phase shift is no greater than 180
approximately 5 measurement points , interpolated error- correction o ers a great improvement
over uncorrected measurements . The accuracy of interpolated error-correction improves as
the phase shift and amplitude change between adjacent points decrease . When you use the
analyzer in linear frequency sweep , perform the original calibration with at least 67 points per
1 GHz of frequency span for greatest accuracy with interpolated error-correction.
Interpolated error-correction is available in three sweep modes: linear frequency , power sweep ,
and CW time .
If there is a valid correction array for a linear frequency sweep , this may be
interpolated to provide correction at the CW frequency used in power sweep
or CW time modes . This correction is part of the interpolated error-correction
feature and is not speci ed.
6-72 Application and Operation Concepts
Interpolated Error-correction · · per

There are twelve di erent error terms for a two-port measurement that can be corrected
by accuracy enhancement in the analyzer . These are directivity , source match, load match,
isolation, re ection tracking, and transmission tracking, each in both the forward and reverse
direction. The analyzer has several di erent measurement calibration routines to characterize
one or more of the systematic error terms and remove their e ects from the measured data.
The calibrate menu allows you to perform the measurement calibration routines . These
procedures range from a simple frequency response calibration to a full two-port calibration
that e ectively removes all twelve error terms .
The response calibration, activated by pressing the
RESPONSE softkey within the calibrate
menu, provides a normalization of the test setup for re ection or transmission measurements .
This calibration procedure may be adequate for measurement of well matched devices . This
is the simplest error-correction to perform, and should be used when extreme measurement
accuracy is not required.
Response and Isolation Calibration
The response and isolation calibration, activated by pressing the
RESPONSE & ISOL'N softkey
within the calibrate menu, provides a normalization for frequency response and crosstalk
errors in transmission measurements , or frequency response and directivity errors in re ection
measurements . This procedure may be adequate for measurement of well matched high-loss
devices .
one-port calibration procedures , activated by pressing the
S11 1-PORT
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
S22 1-PORT softkey within the calibrate menu, provide directivity , source match, and
frequency response vector error-correction for re ection measurements . These procedures
provide high accuracy re ection measurements of one-port devices or properly terminated
two-port devices .
Full Two-P ort Calibration
The full two-port calibration, activated by pressing the
FULL 2-PORT softkey within the
calibrate menu, provides directivity , source match, load match, isolation, and frequency
response vector error-correction, in both forward and reverse directions , for transmission and
re ection measurements of two-port devices . This calibration provides the best magnitude and
phase measurement accuracy for both transmission and re ection measurements of two-port
devices , and requires an S-parameter test set.
In this type of calibration, both forward and reverse measurements must be made . Y ou have
the option of setting the ratio of the number of forward (or reverse) sweeps versus the number
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
of reverse (or forward) sweeps . T o access this function, press
MORE
enter the number of sweeps desired.
Application and Operation Concepts 6-73

TRL*/LRM* Two-P ort Calibration
The TRL*/LRM* two-port calibration, activated by pressing the
within the calibrate menu, provides the ability to make calibrations using the TRL or LRM
method. F or more information, refer to \TRL*/LRM* Calibration," located later in this section.
6-74 Application and Operation Concepts
TRL*/LRM* 2-PORT softkey

If you interrupt a calibration to go to another menu, such as averaging, you can continue the
calibration by pressing the
RESUME CAL SEQUENCE softkey in the correction menu.
The cal kit menu provides access to a series of menus used to specify the characteristics of
calibration standards . The following softkeys are located within the cal kit menu:
SELECT CAL KIT
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SAVE USER KIT
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MODIFY [ ]
The Select Cal Kit Menu
SELECT CAL KIT softkey within the cal kit menu provides access to the select
cal kit menu. This menu allows you to select from several default calibration kits that have
di erent connector types . These kits have prede ned standards and are valid for most
applications . It is not possible to overwrite these standard de nitions .
The numerical de nitions for most Hewlett-P ackard calibration kits can be referenced in the
calibration kit operating and service manuals , or can be viewed on the analyzer . The standard
de nitions can also be modi ed to any set of standards used.
Application and Operation Concepts 6-75
Restarting a Calibration · Cal Kit Menu · Pressing the

Modifying calibration kits is necessary only if unusual standards (such as in TRL*) are used
or the very highest accuracy is required. Unless a calibration kit model is provided with the
calibration devices used, a solid understanding of error-correction and the system error model
are absolutely essential to making modi cations . Y ou may use modi cations to a prede ned
calibration kit by modifying the kit and saving it as a user kit. The original prede ned
calibration kit will remain unchanged.
Before attempting to modify calibration standard de nitions , you should read application note
8510-5A to improve your understanding of modifying calibration kits . The part number of this
application note is 5956-4352. Although the application note is written for the HP 8510 family
of network analyzers , it also applies to the HP 8753D .
Several situations exist that may require a user-de ned calibration kit:
A calibration is required for a connector interface di erent from the four default calibration
A calibration with standards (or combinations of standards) that are di erent from the
default calibration kits is required. (Example: Using three o set shorts instead of open,
short, and load to perform a 1-port calibration.)
The built-in standard models for default calibration kits can be improved or re ned.
Remember that the more closely the model describes the actual performance of the standard,
of 50.0 ohms .)
The following are de nitions of terms:
A \standard" (represented by a number 1-8) is a speci c , well-de ned, physical device used
to determine systematic errors . F or example , standard 1 is a short in the 3.5 mm calibration
kit. Standards are assigned to the instrument softkeys as part of a class .
A standard \type" is one of ve basic types that de ne the form or structure of the model to
be used with that standard (short, open, load, delay/thru, and arbitrary impedance); standard
1 is of the type short in the 3.5 mm calibration kit.
Standard \coecients" are numerical characteristics of the standards used in the model
selected. F or example , the o set delay of the short is 32 ps in the 3.5 mm calibration kit.
A standard \class" is a grouping of one or more standards that determines which of the eight
standards are used at each step of the calibration. F or example , standard number 2 and 8
usually makes up the S
A re ection class , which for type-N calibration kits are male and
female shorts .
The following steps are used to modify or de ne a user kit:
kit.)
a. De ne which \type" of standard it is .
b. De ne the electrical characteristics (coecients) of the standard.
6-76 Application and Operation Concepts
Modifying Calibration Kits · De nitions · Procedure

F or a step by step procedure on how to modify calibration kits , refer to \Modifying Calibration
Kit Standards" located in Chapter 5, \Optimizing Measurement Results ."
Modify Calibration Kit Menu
MODIFY [ ] softkey in the cal kit menu provides access to the modify calibration kit
menu. This leads in turn to additional series of menus associated with modifying calibration
kits . The following is a description of the softkeys located within this menu:
DEFINE STANDARD makes the standard number the active function, and brings up the
de ne standard menus . Before selecting a standard, a standard number must be entered.
This number (1 to 8) is an arbitrary reference number used to reference standards while
specifying a class . The standard numbers for the prede ned calibration kits are as follows:
Although the numbering sequences are arbitrary , confusion can be minimized
by using consistency . However , standard 5 is always a sliding load.
SPECIFY CLASS leads to the specify class menu. After the standards are modi ed, use this
key to specify a class to group certain standards .
LABEL CLASS leads to the label class menu, to give the class a meaningful label for future
reference .
LABEL KIT leads to a menu for constructing a label for the user-modi ed cal kit. If a label
is supplied, it will appear as one of the ve softkey choices in the select cal kit menu. The
approach is similar to de ning a display title , except that the kit label is limited to ten
characters .
TRL/LRM OPTION brings up the TRL Option menu.
KIT DONE (MODIFIED) terminates the calibration kit modi cation process , after all
standards are de ned and all classes are speci ed. Be sure to save the kit with the
SAVE USER KIT softkey , if it is to be used later .
Application and Operation Concepts 6-77

De ne Standard Menus
Standard de nition is the process of mathematically modeling the electrical characteristics
(delay , attenuation, and impedance) of each calibration standard. These electrical
characteristics (coecients) can be mathematically derived from the physical dimensions and
material of each calibration standard, or from its actual measured response . The parameters of
the standards can be listed in T able 6-4.
T able 6-4. Standard De nitions
= Calibration Kit Label:

of network analyzer is set to this value .
Open, short, load, delay/thru, or arbitrary impedance .
Load or arbitrary impedance only .
Arbitrary impedance only , device terminating impedance .
Open standard types only .
6-78 Application and Operation Concepts

Each standard must be identi ed as one of ve \types": open, short, load, delay/thru, or
arbitrary impedance .
After a standard number is entered, selection of the standard type will present one of ve
menus for entering the electrical characteristics (model coecients) corresponding to that
standard type , such as
OPEN . These menus are tailored to the current type , so that only
characteristics applicable to the standard type can be modi ed.
The following is a description of the softkeys located within the de ne standard menu:
OPEN de nes the standard type as an open, used for calibrating re ection measurements .
Opens are assigned a terminal impedance of in nite ohms , but delay and loss o sets may
still be added. Pressing this key also brings up a menu for de ning the open, including its
capacitance .
As a re ection standard, an open termination o ers the advantage of broadband frequency
coverage . At microwave frequencies , however , an open rarely has perfect re ection
characteristics because fringing capacitance e ects cause phase shift that varies with
frequency . This can be observed in measuring an open termination after calibration, when
an arc in the lower right circumference of the Smith chart indicates capacitive reactance .
These e ects are impossible to eliminate , but the calibration kit models include the open
termination capacitance at all frequencies for compatible calibration kits . The capacitance
model is a cubic polynomial, as a function of frequency , where the polynomial coecients
are user-de nable . The capacitance model equation is:
C = (C0) + (C1 F) + (C2 F
where F is the measurement frequency .
The terms in the equation are de ned with the specify open menu as follows:
C0 allows you to enter the C0 term, which is the constant term of the cubic polynomial
and is scaled by 10
F arads .
C1 allows you to enter the C1 term, expressed in F/Hz (F arads/Hz) and scaled by 10
C2 allows you to enter the C2 term, expressed in F/Hz
and scaled by 10
NNNNNNN
C3 allows you to enter the C3 term, expressed in F/Hz
and scaled by 10
SHORT de nes the standard type as a short, for calibrating re ection measurements . Shorts
are assigned a terminal impedance of 0 ohms , but delay and loss o sets may still be added.
NNNNNNNNNNNNN
LOAD de nes the standard type as a load (termination). Loads are assigned a terminal
impedance equal to the system characteristic impedance Z0, but delay and loss o sets
may still be added. If the load impedance is not Z0, use the arbitrary impedance standard
de nition.
FIXED de nes the load as a xed (not sliding) load.
SLIDING de nes the load as a sliding load. When such a load is measured during
calibration, the analyzer will prompt for several load positions , and calculate the ideal load
value from it.
OFFSET de nes the load as being o set.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DELAY/THRU de nes the standard type as a transmission line of speci ed length, for
calibrating transmission measurements .
Application and Operation Concepts 6-79

ARBITRARY IMPEDANCE de nes the standard type to be a load, but with an arbitrary
impedance (di erent from system Z0).
TERMINAL IMPEDANCE allows you to specify the (arbitrary) impedance of the standard, in
ohms .
FIXED de nes the load as a xed (not sliding) load.
SLIDING de nes the load as a sliding load. When such a load is measured during
calibration, the analyzer will prompt for several load positions , and calculate the ideal load
value from it.
Any standard type can be further de ned with o sets in delay , loss , and standard impedance;
assigned minimum or maximum frequencies over which the standard applies; and de ned as
coax or waveguide . The
SPECIFY OFFSET softkey provides access to the specify o set menu
(described next).
LABEL STD softkey allows you to de ne a distinct label for each standard, so that
the analyzer can prompt the user with explicit standard labels during calibration (such as
SHORT ). The function is similar to de ning a display title , except that the label is limited to ten
characters .
After each standard is de ned, including o sets , the
STD DONE (DEFINED) softkey will
terminate the standard de nition.
Specify O set Menu
The specify o set menu allows additional speci cations for a user-de ned standard. F eatures
speci ed in this menu are common to all ve types of standards .
O sets may be speci ed with any standard type . This means de ning a uniform length of
transmission line to exist between the standard being de ned and the actual measurement
plane . (Example: a waveguide short circuit terminator , o set by a short length of waveguide .)
F or re ection standards , the o set is assumed to be between the measurement plane and the
standard (one-way only). F or transmission standards , the o set is assumed to exist between the
two reference planes (in e ect, the o set is the thru). Three characteristics of the o set can
be de ned: its delay (length), loss , and impedance .
In addition, the frequency range over which a particular standard is valid can be de ned with
a minimum and maximum frequency . This is particularly important for a waveguide standard,
since its behavior changes rapidly beyond its cuto frequency . Note that several band-limited
standards can together be de ned as the same \class" (see specify class menu). Then, if a
measurement calibration is performed over a frequency range exceeding a single standard,
additional standards can be used for each portion of the range .
Lastly , the standard must be de ned as either coaxial or waveguide . If it is waveguide ,
dispersion e ects are calculated automatically and included in the standard model.
The following is a description of the softkeys located within the specify o set menu:
OFFSET DELAY allows you to specify the one-way electrical delay from the measurement
(reference) plane to the standard, in seconds (s). (In a transmission standard, o set delay is
the delay from plane to plane .) Delay can be calculated from the precise physical length of
the o set, the permittivity constant of the medium, and the speed of light.
In coax, group delay is considered constant. In waveguide , however , group delay is
dispersive , that is , it changes signi cantly as a function of frequency . Hence , for a waveguide
standard, o set delay must be de ned as though it were a TEM wave (without dispersion).
6-80 Application and Operation Concepts
The

OFFSET LOSS allows you to specify energy loss , due to skin e ect, along a one-way length of
coax o set. The value of loss is entered as ohms/nanosecond (or Giga ohms/second) at 1 GHz.
(Such losses are negligible in waveguide , so enter 0 as the loss o set.)
NNNNNNNNNNNNNNNNNNNNNNNNNNNN
OFFSET Z0 allows you to specify the characteristic impedance of the coax o set. (Note:
This is not the impedance of the standard itself.) F or waveguide , the o set impedance as
well as the system Z0 must always be set to 1 .
MINIMUM FREQUENCY allows you to de ne the lowest frequency at which the standard
can be used during measurement calibration. In waveguide , this must be the lower cuto
frequency of the standard, so that the analyzer can calculate dispersive e ects correctly (see
OFFSET DELAY above).
MAXIMUM FREQUENCY allows you to de ne the highest frequency at which the standard can
be used during measurement calibration. In waveguide , this is normally the upper cuto
frequency of the standard.
COAX de nes the standard (and the o set) as coaxial. This causes the analyzer to assume
linear phase response in any o sets .
WAVEGUIDE de nes the standard (and the o set) as rectangular waveguide . This causes the
analyzer to assume a dispersive delay (see
OFFSET DELAY above).
This menu allows you to label (reference) individual standards during the menu-driven
measurement calibration sequence . The labels are user-de nable using a character set shown
on the display that includes letters , numbers , and some symbols , and they may be up to ten
characters long. The analyzer will prompt you to connect standards using these labels , so they
should be meaningful to you, and distinct for each standard.
By convention, when sexed connector standards are labeled male (m) or female (f), the
designation refers to the test port connector sex, not the connector sex of the standard.
Once a standard has been de ned, it must be assigned to a standard \class ." This is a group of
from one to seven standards that is required to calibrate for a single error term. The standards
within a single class can be assigned to the locations listed in T able 6-5 according to their
standard reference numbers .
A class often consists of a single standard, but may be composed of more than one standard if
band-limited standards are used. F or example , if there were two load standards a xed load
for low frequencies , and a sliding load for high frequencies then that class would have two
standards .
Application and Operation Concepts 6-81
Label Standard Menu · Specify Class Menu

T able 6-5. Standard Class Assignments
Calibration Kit Label:

The number of standard classes required depends on the type of calibration being performed,
and is identical to the number of error terms corrected. A response calibration requires only
one class , and the standards for that class may include an open, or short, or thru. A 1-port
calibration requires three classes . A full 2-port calibration requires 10 classes , not including two
for isolation.
The number of standards that can be assigned to a given class may vary from none (class not
used) to one (simplest class) to seven. When a certain class of standards is required during
calibration, the analyzer will display the labels for all the standards in that class (except when
the class consists of a single standard). This does not, however , mean that all standards in a
class must be measured during calibration. Unless band-limited standards are used, only a
single standard per class is required.
It is often simpler to keep the number of standards per class to the bare
minimum needed (often one) to avoid confusion during calibration.
Each class can be given a user-de nable label as described under label class menus .
Standards are assigned to a class simply by entering the standard's reference number
(established while de ning a standard) under a particular class . The following is a description
of the softkeys located within the specify class menu:
S11A allows you to enter the standard numbers for the rst class required for an S
calibration. (F or default calibration kits , this is the open.)
6-82 Application and Operation Concepts
1-port

S11B allows you to enter the standard numbers for the second class required for an S
1-port calibration. (F or default calibration kits , this is the short.)
S11C allows you to enter the standard numbers for the third class required for an S
calibration. (F or default calibration kits , this is the load.)
S22A allows you to enter the standard numbers for the rst class required for an S
calibration. (F or default calibration kits , this is the open.)
S22B allows you to enter the standard numbers for the second class required for an S
1-port calibration. (F or default calibration kits , this is the short.)
S22C allows you to enter the standard numbers for the third class required for an S
calibration. (F or default calibration kits , this is the load.)
FWD TRANS. allows you to enter the standard numbers for the forward transmission thru
calibration. (F or default calibration kits , this is the thru.)
REV TRANS. allows you to enter the standard numbers for the reverse transmission (thru)
calibration. (F or default calibration kits , this is the thru.)
FWD MATCH allows you to enter the standard numbers for the forward match (thru)
calibration. (F or default calibration kits , this is the thru.)
REV MATCH allows you to enter the standard numbers for the reverse match (thru)
calibration. (F or default calibration kits , this is the thru.)
RESPONSE allows you to enter the standard numbers for a response calibration. This
calibration corrects for frequency response in either re ection or transmission measurements ,
depending on the parameter being measured when a calibration is performed. (F or default
kits , the standard is either the open or short for re ection measurements , or the thru for
transmission measurements .)
RESPONSE & ISOL'N allows you to enter the standard numbers for a response & isolation
calibration. This calibration corrects for frequency response and directivity in re ection
measurements , or frequency response and isolation in transmission measurements .
NNNNNNNNNNNNNNNNNNNNNNNNN
TRL THRU allows you to enter the standard numbers for a TRL thru calibration.
TRL REFLECT allows you to enter the standard numbers for a TRL re ect calibration.
TRL LINE OR MATCH allows you to enter the standard numbers for a TRL line or match
calibration.
The label class menus are used to de ne meaningful labels for the calibration classes . These
then become softkey labels during a measurement calibration. Labels can be up to ten
characters long.
LABEL KIT softkey within the modify cal kit menu, provides access to this menu. It
is identical to the label class menu and the label standard menu described above . It allows
de nition of a label up to eight characters long.
After a new calibration kit has been de ned, be sure to specify a label for it. Choose a label
that describes the connector type of the calibration devices . This label will then appear in the
Application and Operation Concepts 6-83

CAL KIT [ ] softkey label in the correction menu and the
MODIFY [ ] label in the select
cal kit menu. It will be saved with calibration sets .
Once a measurement calibration has been generated with a user-de ned calibration kit, its
performance should be checked before making device measurements . T o check the accuracy
that can be obtained using the new calibration kit, a device with a well-de ned frequency
response (preferably unlike any of the standards used) should be measured. The veri cation
device must not be one of the calibration standards: measurement of one of these standards is
merely a measure of repeatability .
T o achieve more complete veri cation of a particular measurement calibration, accurately
known veri cation standards with a diverse magnitude and phase response should be
used. National standard traceable or HP standards are recommended to achieve veri able
measurement accuracy .
The published speci cations for the HP 8753D network analyzer system
include accuracy enhancement with compatible calibration kits . Measurement
calibrations made with user-de ned or modi ed calibration kits are not subject
to the HP 8753D speci cations , although a procedure similar to the system
veri cation procedure may be used.
6-84 Application and Operation Concepts
V erify performance

The HP 8753D RF network analyzer has the capability of making calibrations using the \TRL"
(thru-re ect-line) method. This section contains information on the following subjects:
Why Use TRL Calibration?
How TRL*/LRM* Calibration W orks
Improving Raw Source Match and Load Match F or TRL*/LRM* Calibration
The TRL Calibration Procedure
Requirements for TRL Standards
Why Use TRL Calibration?
This method is convenient in that calibration standards can be fabricated for a speci c
measurement environment, such as a transistor test xture or microstrip . Microstrip devices
in the form of chips , MMIC's , packaged transistors , or beam-lead diodes cannot be connected
directly to the coaxial ports of the analyzer . The device under test (DUT) must be physically
connected to the network analyzer by some kind of transition network or xture . Calibration
for a xtured measurement in microstrip presents additional diculties .
A calibration at the coaxial ports of the network analyzer removes the e ects of the network
analyzer and any cables or adapters before the xture; however , the e ects of the xture itself
are not accounted for . An in- xture calibration is preferable , but high-quality Short-Open-Load-
Thru (SOL T) standards are not readily available to allow a conventional Full 2-port calibration
of the system at the desired measurement plane of the device . In microstrip , a short circuit is
inductive , an open circuit radiates energy , and a high-quality purely resistive load is dicult to
produce over a broad frequency range . The Thru-Re ect-Line (TRL) 2-port calibration is an
alternative to the traditional SOL T Full 2-port calibration technique that utilizes simpler , more
convenient standards for device measurements in the microstrip environment.
F or coaxial, waveguide and other environments where high-quality impedance standards are
readily available , the traditional short, open, load, thru (SOL T) method provides the most
accurate results since all of the signi cant systematic errors are reduced. This method is
implemented in the form of the S
1-port, and full 2-port calibration selections .
In all measurement environments , the user must provide calibration standards for the desired
calibration to be performed. The advantage of TRL is that only three standards need to
be characterized as opposed to 4 in the traditional open, short, load, and thru full 2-port
calibrations . Further , the requirements for characterizing the T , R, and L standards are less
stringent and these standards are more easily fabricated.
standards used. F or example , \LRL" indicates that two lines and a re ect standard are used;
\TRM" indicates that a thru, re ection and match standards are used. All of these refer to the
same basic method.
Application and Operation Concepts 6-85

How TRL*/LRM* Calibration W orks
The TRL*/LRM* calibration used in the HP 8753D relies on the characteristic impedance
of simple transmission lines rather than on a set of discrete impedance standards . Since
transmission lines are relatively easy to fabricate (in a microstrip , for example), the impedance
of these lines can be determined from the physical dimensions and substrate's dielectric
constant.

Figure 6-51.
HP 8753D functional block diagram for a 2-port error-corrected
F or an HP 8753D TRL* 2-port calibration, a total of 10 measurements are made to quantify
eight unknowns (not including the two isolation error terms). Assume the two transmission
leakage terms , E
, are measured using the conventional technique . The eight TRL
XR
error terms are represented by the error adapters shown in Figure 6-52. Although this error
model is slightly di erent from the traditional Full 2-port 12-term model, the conventional
error terms may be derived from it. F or example , the forward re ection tracking (E
represented by the product of "
. Also notice that the forward source match (E
reverse load match (E
) are both represented by "
, while the reverse source match (E
SR
and forward load match (E
) are both represented by "
. In order to solve for these eight
unknown TRL error terms , eight linearly independent equations are required.
The rst step in the TRL* 2-port calibration process is the same as the transmission step for
a Full 2-port calibration. F or the thru step , the test ports are connected together directly
(zero length thru) or with a short length of transmission line (non- zero length thru) and the
transmission frequency response and port match are measured in both directions by measuring
all four S-parameters .
F or the re ect step , identical high re ection coecient standards (typically open or short
circuits) are connected to each test port and measured (S
).
F or the line step , a short length of transmission line (di erent in length from the thru) is
inserted between port 1 and port 2 and again the frequency response and port match are
measured in both directions by measuring all four S-parameters .
In total, ten measurements are made , resulting in ten independent equations . However , the
TRL error model has only eight error terms to solve for . The characteristic impedance of the
6-86 Application and Operation Concepts

line standard becomes the measurement reference and, therefore , has to be assumed ideal (or
known and de ned precisely).
At this point, the forward and reverse directivity (E
DR
), transmission tracking (E
TR
), and re ection tracking (E
RR
) terms may be derived from the TRL error
terms . This leaves the isolation (E
XR
), source match (E
SR
) and load match
LF
LR
) terms to discuss .
Two additional measurements are required to solve for the isolation terms (E
XR
).
Isolation is characterized in the same manner as the Full 2-port calibration. F orward and
reverse isolation are measured as the leakage (or crosstalk) from port 1 to port 2 with each port
terminated. The isolation part of the calibration is generally only necessary when measuring
high loss devices (greater than 70 dB).
If an isolation calibration is performed, the xture leakage must be the
same during the isolation calibration and the measurement.

Figure 6-52. 8-term TRL error model and generalized coecients
Application and Operation Concepts 6-87

Source match and load match
A TRL calibration assumes a perfectly balanced test set architecture as shown by the term
which represents both the forward source match (E
) and reverse load match (E
) and by the
term which represents both the reverse source match (E
) and forward load match (E
).
However , in any switching test set, the source and load match terms are not equal because the
transfer switch presents a di erent terminating impedance as it is changed between port 1 and
port 2.
Because the standard HP 8753D network analyzer is based on a three-sampler receiver
architecture , it is not possible to di erentiate the source match from the load match terms . The
terminating impedance of the switch is assumed to be the same in either direction. Therefore ,
the test port mismatch cannot be fully corrected. An assumption is made that:
forward source match (E
) = reverse load match (E
reverse source match (E
) = forward load match (E
F or a xture , TRL* can eliminate the e ects of the xture's loss and length, but does not
completely remove the e ects due to the mismatch of the xture .
Because the TRL technique relies on the characteristic impedance of
transmission lines , the mathematically equivalent method LRM (for
line-re ect-match) may be substituted for TRL. Since a well matched
termination is , in essence , an in nitely long transmission line , it is well suited
for low (RF) frequency calibrations . A chieving a long line standard for low
frequencies is often times physically impossible .
Improving Raw Source Match and Load Match F or TRL*/LRM* Calibration
A technique that can be used to improve the raw test port mismatch is to add high quality
xed attenuators . The e ective match of the system is improved because the xed attenuators
usually have a return loss that is better than that of the network analyzer . A dditionally , the
attenuators provide some isolation of re ected signals . The attenuators also help to minimize
the di erence between the port source match and load match, making the error terms more
equivalent.
With the attenuators in place , the e ective port match of the system is improved so that the
mismatch of the xture transition itself dominates the measurement errors after a calibration.
6-88 Application and Operation Concepts
" · ) = " · ) = "


Figure 6-53. Typical Measurement Set up
If the device measurement requires bias , it will be necessary to add external bias tees between
the xed attenuators and the xture . The internal bias tees of the analyzer will not pass the
bias properly through the external xed attenuators . Be sure to calibrate with the external bias
tees in place (no bias applied during calibration) to remove their e ect from the measurement.
Because the bias tees must be placed after the attenuators , they essentially become part of the
xture . Therefore , their mismatch e ects on the measurement will not be improved by the
attenuators .
Although the xed attenuators improve the raw mismatch of the network analyzer system,
they also degrade the overall measurement dynamic range .
This e ective mismatch of the system after calibration has the biggest e ect on re ection
measurements of highly re ective devices . Likewise , for well matched devices , the e ects of
mismatch are negligible . This can be shown by the following approximation:
Re ection magnitude uncertainty = E
S
+ E
(S
+ E
S
S
D
R
S
Transmission magnitude uncertainty = E
S
+ E
S
S
+ E
S
S
T
S
where:
= e ective directivity
E
= e ective re ection tracking
E
= e ective source match
E
= e ective load match
= e ective crosstalk
= e ective transmission tracking
S
=S-parameters of the device under test
Application and Operation Concepts 6-89
)

The TRL Calibration Procedure
Requirements for TRL Standards
When building a set of TRL standards for a microstrip or xture environment, the requirements
for each of these standard types must be satis ed.
No loss . Characteristic impedance (Z
) need not be known.
= S
S
= S
THRU
Z
of the thru must be the same as the line . (If they are not the same , the
average impedance is used.)
Attenuation of the thru need not be known.
If the thru is used to set the reference plane , the insertion phase or
electrical length must be well-known and speci ed. If a non-zero length
thru is speci ed to have zero delay , the reference plane is established in the
middle of the thru.
REFLECT Re ection coecient ( ) magnitude is optimally 1.0, but need not be
known.
Phase of must known and speci ed to within 1/4 wavelength or 90
During computation of the error model, the root choice in the solution of a
quadratic equation is based on the re ection data. An error in de nition
would show up as a 180
error in the measured phase .
must be identical on both ports .
If the re ect is used to set the reference plane , the phase response must be
well-known and speci ed.
Z
of the line establishes the reference impedance of the measurement
(i.e . S
= 0). The calibration impedance is de ned to be the same as
of the line . If the Z
is known but not the desired value (i.e ., not equal to
50 ), the SYSTEMS Z
selection under the TRL/LRM options menu is used.
Insertion phase of the line must not be the same as the thru (zero length
or non-zero length). The di erence between the thru and line must be
) n x 180
. Measurement uncertainty will increase
signi cantly when the insertion phase nears 0 or an integer multiple of 180
Optimal line length is 1/4 wavelength or 90
of insertion phase relative to
the thru at the middle of the desired frequency span.
Usable bandwidth for a single thru/line pair is 8:1 (frequency span:start
frequency).
Multiple thru/line pairs (Z
assumed identical) can be used to extend the
bandwidth to the extent transmission lines are available .
Attenuation of the line need not be known.
Insertion phase must be known and speci ed within
1/4 wavelength or 90
Z
of the match establishes the reference impedance of the measurement.
must be identical on both ports .
6-90 Application and Operation Concepts

F abricating and de ning calibration standards for TRL/LRM
When calibrating a network analyzer , the actual calibration standards must have known
physical characteristics . F or the re ect standard, these characteristics include the o set in
electrical delay (seconds) and the loss (ohms/second of delay). The characteristic impedance ,
OFFSET Z0 , is not used in the calculations in that it is determined by the line standard. The
re ection coecient magnitude should optimally be 1.0, but need not be known since the same
re ection coecient magnitude must be applied to both ports .
The thru standard may be a zero-length or known length of transmission line . The value of
length must be converted to electrical delay , just like that done for the re ect standard. The
loss term must also be speci ed.
Application and Operation Concepts 6-91

The line standard must meet speci c frequency related criteria, in conjunction with the length
used by the thru standard. In particular , the insertion phase of the line must not be the same
as the thru. The optimal line length is 1/4 wavelength (90 degrees) relative to a zero length
thru at the center frequency of interest, and between 20 and 160 degrees of phase di erence
over the frequency range of interest. (Note: these phase values can be N 180 degrees
where N is an integer .) If two lines are used (LRL), the di erence in electrical length of
the two lines should meet these optimal conditions . Measurement uncertainty will increase
signi cantly when the insertion phase nears zero or is an integer multiple of 180 degrees , and
this condition is not recommended.
F or a transmission media that exhibits linear phase over the frequency range of interest, the
following expression can be used to determine a suitable line length of one-quarter wavelength
at the center frequency (which equals the sum of the start frequency and stop frequency
divided by 2):
E l ectr ical l eng th (cm) = (LI N E 0 l eng th T H R U )
E l ectr ical l eng th (cm) =
f 1(M H z ) + f 2(M H z )
let:
f1 = 1000 MHz
f2 = 2000 MHz
VF = V elocity F actor = 1 (for this example)
Thus , the length to initially check is 5 cm.
Next, use the following to verify the insertion phase at f1 and f2:
(360 f l )
P hase (deg r ees ) =
where:
l = length of line
v = velocity = speed of light velocity factor
which can be reduced to the following using frequencies in MHz and length in centimeters:
P hase (deg r ees ) appr ox =
So for an air line (velocity factor approximately 1) at 1000 MHz, the insertion phase is
60 degrees for a 5 cm line; it is 120 degrees at 2000 MHz. This line would be a suitable line
standard.
F or microstrip and other fabricated standards , the velocity factor is signi cant. In those cases ,
the phase calculation must be divided by that factor . F or example , if the dielectric constant for
a substrate is 10, and the corresponding \e ective" dielectric constant for microstrip is 6.5,
then the \e ective" velocity factor equals 0.39 (1 square root of 6.5).
Using the rst equation with a velocity factor of 0.39, the initial length to test would be 1
.95 cm. This length provides an insertion phase at 1000 MHz of 60 degrees; at 2000 MHz,
120 degrees (the insertion phase should be the same as the air line because the velocity factor
was accounted for when using the rst equation).
6-92 Application and Operation Concepts
f = frequency

Another reason for showing this example is to point out the potential problem in calibrating at
low frequencies using TRL. F or example , one-quarter wavelength is
Leng th (cm) =
where:
fc = center frequency
Thus , at 50 MHz,
Leng th (cm) =
50 (M H z )
Such a line standard would not only be dicult to fabricate , but its long term stability and
usability would be questionable as well.
Thus at lower frequencies and/or very broad band measurements , fabrication of a \match" or
termination may be deemed more practical. Since a termination is , in essence , an in nitely
long transmission line , it ts the TRL model mathematically , and is sometimes referred to as a
\TRM" calibration.
The TRM calibration technique is related to TRL with the di erence being that it bases the
characteristic impedance of the measurement on a matched Z
termination instead of a
transmission line for the third measurement standard. Like the TRL thru standard, the TRM
THRU standard can either be of zero length or non-zero length. The same rules for thru and
re ect standards used for TRL apply for TRM.
TRM has no inherent frequency coverage limitations which makes it more convenient in some
measurement situations . A dditionally , because TRL requires a di erent physical length for the
thru and the line standards , its use becomes impractical for xtures with contacts that are at a
xed physical distance from each other .
F or information on how to modify calibration constants for TRL*/LRM*, and how to perform a
TRL or TRM calibration, refer to Chapter 5, \Optimizing Measurement Results ."
TRL/LRM OPTION softkey provides access to the TRL/LRM options menu. There are two
selections under this menu:
The characteristic impedance used during the calibration can be referenced to either the line
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
(or match) standard (
defaults to a calibration impedance that is equal to the line (or match) standard.
assumed to match the system impedance exactly (the line standard is re ectionless). After
a calibration, all measurements are referenced to the impedance of the line standard. F or
example , when the line standard is remeasured, the response will appear at the center of the
Smith chart. When
SET ZO (under CAL
OFFSET ZO (within the de ne standard menu) are ignored.
Application and Operation Concepts 6-93

impedance of the line standard. This requires a knowledge of the exact value of the Z
the line . The system reference impedance is set using
SET ZO under the calibration menu.
The actual impedance of the line is set by entering the real part of the line impedance as the
OFFSET ZO within the de ne standard menu. F or example , if the line was known to have
a characteristic impedance of 51 (
OFFSET ZO = 51 ), it could still be used to calibrate
for a 50 measurement (
SET ZO = 50 ). After a calibration, all measurements would be
referenced to 50 , instead of 51 . When the line standard is remeasured, the center of the
Smith chart is at the current value of
SET ZO (in this case , 50 ). Since only one value of
can be selected for the line standard, the value of Z
should be a constant value over
the frequency range of interest in order to be meaningful.
The location of the reference plane is determined by the selection of
must have a known insertion phase or electrical length. If a non-zero length thru is speci ed
to have zero delay , the reference plane will be established in the middle of the thru. The
re ect standard may be used to set the reference plane instead of the thru provided the phase
response (o set delay , reactance values and standard type) of the re ect standard is known
and is speci ed in the calibration kit de nition.
Dispersion occurs when a transmission medium exhibits a variable propagation
or phase velocity as a function of frequency . The result of dispersion is a
non-linear phase shift versus frequency , which leads to a group delay which
is not constant. F ortunately , the TRL calibration technique accounts for
dispersive e ects of the test xture up to the calibration plane , provided that:
and is used to set the reference plane (
SET REF: THRU ).
characteristics (i.e ., identical height, width and relative dielectric constant).
When a non-zero length thru is used to set the reference plane , it should be
de ned as having zero length in the TRL standards de nition, even though it
has physical length. The actual electrical length of the thru standard must then
be subtracted from the actual electrical length of each line standard in the TRL
calibration kit de nition. The device must then be mounted between two short
lengths of transmission line so that each length is exactly one-half of the length
of the non-zero length thru standard. In this con guration, the measurement
will be properly calibrated up to the point of the device .
6-94 Application and Operation Concepts
of · o set Z · Dispersion E ects

P ower Meter Calibration
PWRMTR CAL [ ] softkey within the correction menu, leads to a series of menus
associated with power meter calibration.
An HP-IB-compatible power meter can monitor and correct RF source power to achieve leveled
power at the test port. During a power meter calibration, the power meter samples the power
at each measurement point across the frequency band of interest. The analyzer then constructs
a correction data table to correct the power output of the internal source . The correction table
may be saved in an instrument state register with the
SAVE key .
The correction table may be updated on each sweep (in a leveling application) or during
an initial single sweep . In the sample-and-sweep mode the power meter is not needed for
subsequent sweeps . The correction table may be read or modi ed through HP-IB .
when you are testing a system with signi cant frequency response errors ( F or example , a
coupler with signi cant roll-o , or a long cable with a signi cant amount of loss .)
when you are measuring devices that are very sensitive to actual input power for proper
when you require a reference for receiver power calibration
Calibrated P ower Level
By setting the analyzer calibrated power to the desired value at the power meter , this power
level will be maintained at that port during the entire sweep . First set the source power so that
the power at the test device is approximately correct. This reduces residual power errors when
only one reading is taken. Refer to
NUMBER OF READINGS softkey description in
Chapter 9, \Key De nitions ." When power meter calibration is on, the annotation PC is
displayed. This indicates that the source power is being updated during the sweep . Calibrated
power level becomes the active entry if any of the following softkeys are pressed:
PWRMTR CAL [OFF]
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
EACH SWEEP
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ONE SWEEP
NNNNNNNNNNNNNNNNN
POWER (if power meter cal is on)
Regardless of the measurement application, the analyzer's source can only supply corrected
power within the selected power range . If power outside this range is requested, the
annotation will change to PC? .
P ower meter calibration may be used in linear , log, list, CW , and power sweep modes . In power
sweep , the power at each point is the true power at the power meter , not the power at the
analyzer's source output.
Application and Operation Concepts 6-95

Loss of P ower Meter Calibration Data
The power meter calibration data will be lost by committing any of the following actions:
Turning power o . Turning o the instrument erases the power meter calibration table .
Changing sweep type . If the sweep type is changed (linear , log, list, CW , power) while
power meter calibration is on, the calibration data will be lost. However , calibration data is
retained if you change the sweep type while power meter calibration is o .
Changing frequency . P ower meter calibration data will also be lost if the frequency is
changed in log or list mode , but it is retained in linear sweep mode .
. Presetting the instrument will erase power meter calibration data. If the
instrument state has been saved in a register using the
key , you may recall the
instrument state and the data will be restored. Saving the instrument state will not protect
the data if the instrument is turned o .
Interpolation in P ower Meter Calibration
If the frequency is changed in linear sweep , or the start/stop power is changed in power sweep ,
then the calibration data is interpolated for the new range .
If calibration power is changed in any of the sweep types , the values in the power setting array
are increased or decreased to re ect the new power level. Some accuracy is lost when this
occurs .
P ower Meter Calibration Modes of Operation
Continuous Sample Mode (Each Sweep)
Y ou can set the analyzer to update the correction table at each point for sweep (as in a
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
leveling application), using the
EACH SWEEP softkey . In this mode , the analyzer checks the
power level at every frequency point each time it sweeps . Y ou can also have more than one
sample/correction iteration at each frequency point. (See the
NUMBER OF READINGS softkey
description in Chapter 9.)
While using the continuous sample mode , the power meter must remain connected as shown
in Figure 6-54. A power splitter or directional coupler samples the actual power going to the
test device and is measured by the power meter . The power meter measurement provides the
information necessary to update the correction table via HP-IB .
Continuous correction slows the sweep speed considerably , especially when low power levels
are being measured by the power meter . It may take up to 10 seconds per point if the power
level is less than 20 dBm. F or faster operation, you can use the sample-and-sweep mode . If
you use a directional coupler , you must enter the attenuation of the coupled arm with respect
to the through arm using the
POWER LOSS softkey .
6-96 Application and Operation Concepts
Pressing

Figure 6-54. T est Setup for Continuous Sample Mode
Sample-and-Sweep Mode (One Sweep)
Y ou can use the
ONE SWEEP softkey to activate the sample-and-sweep mode . This will correct
the analyzer output power and update the power meter calibration data table during the initial
measurement sweep . In this mode of operation, the analyzer does not require the power meter
for subsequent sweeps . Y ou may use a power splitter or directional coupler , or simply connect
the power sensor directly to the analyzer to measure the power for the initial sweep prior to
connecting and measuring the test device (see Figure 6-55).
The speed of the calibration will be slow while power meter readings are taken (see T able 6-6 ).
However , once the sample sweep is nished, subsequent sweeps are power-corrected using
the data table , and sweep speed increases signi cantly . Once the initial sweep is taken,
sample-and-sweep correction is much faster than continuous sample correction.
If the calibrated power level is changed after the initial measurement sweep is done , the entire
correction table is increased or decreased by that amount and the annotation PCo (indicating
power calibration o set) appears on the display . The resulting power will no longer be as
accurate as the original calibration.
Figure 6-55. T est Setup for Sample-and-Sweep Mode
Application and Operation Concepts 6-97

P ower Loss Correction List
If a directional coupler or power splitter is used to sample the RF power output of the analyzer ,
the RF signal going to the power meter may be di erent than that going to the test device . A
directional coupler will attenuate the RF signal by its speci ed coupling factor . The di erence
in attenuation between the through arm and the coupled arm (coupling factor) must be entered
using the loss/sensor list menu. Non-linearities in either the directional coupler or power
splitter can be corrected in the same way .
P ower loss information is entered in much the same way as limit line parameters . Up to 12
segments may be entered, each with a di erent attenuation value . The entered data will not
be lost if the instrument's power is cycled.
P ower Sensor Calibration F actor List
Two power sensor calibration data lists can be created in the analyzer . No single power sensor
covers the entire frequency range of the analyzer , therefore the calibration data for two
di erent power sensors must be available . The entered data will not be lost if the instrument's
power is cycled.
Speed and A ccuracy
The speed and accuracy of a power meter calibration vary depending on the test setup and the
measurement parameters . F or example , number of points , number of readings , if the power is
less than 20 dBm, continuous versus sample and sweep mode . A ccuracy is improved if you
set the source power such that it is approximately correct at the measurement port. P ower
meter calibration should then be turned on. With number of readings = 2, very accurate
measurements are achieved.
T able 6-6 shows typical sweep speed and power accuracy . The times given apply only to the
test setup for continuous correction or for the rst sweep of sample-and-sweep correction.
The typical values given in T able 6-6 were derived under the following conditions:
T est Equipment Used
HP 8753D network analyzer
HP 436A power meter
HP 8485A power sensor
The time required to perform a power meter calibration depends on the source power and
number of points tested. The parameters used to derive the typical values in T able 6-6 are as
follows:
Sweep time is linearly proportional to the number of points measured. F or example , a sweep
taking 49 seconds at 51 points will take approximately 98 seconds if 101 points are measured.
6-98 Application and Operation Concepts
Stimulus P arameters

T able 6-6.
Characteristic P ower Meter Calibration Sweep Speed and A ccuracy
Number of Readings Sweep Time
at T est P ort (dBm)
1 33 0.7
+5 2 64 0.2
3 95 0.1
1 48 0.7
15 2 92 0.2
3 123 0.1
1 194 0.7
30 2 360 0.2
3 447 0.1
1 Sweep speed applies to every sweep in continuous correction mode , and
to the rst sweep in sample-and-sweep mode . Subsequent sweeps in
sample-and-sweep mode will be much faster .
2 The accuracy values were derived by combining the accuracy of the
power meter and linearity of the analyzer's internal source , as well as
the mismatch uncertainty associated with the power sensor .
The accuracy values in T able 6-6 were derived by combining the accuracy of the power meter
and linearity of the analyzer's internal source , as well as the mismatch uncertainty associated
with the test set and the power sensor .
P ower meter calibration measures the source power output (at the measurement port) at a
single stimulus point, and compares it to the calibrated power you selected. If the two values
are di erent, power meter calibration changes the source output power by the di erence . This
process is repeated at every stimulus point. The accuracy of the result depends on the amount
of correction required. If the selected number of readings = 1, the nal measurement accuracy
is signi cantly a ected by a large power change . However , if the selected number of readings
is >1, the power change on the second or third reading is much smaller: thus accuracy is much
better .
Set source power approximately correct at the measurement port, then activate power meter
calibration. This method can signi cantly increase the accuracy of the measurement when the
selected number of readings = 1. Smaller accuracy improvements occur with a higher number
of readings . Remember that mismatch errors a ect accuracy as well.
P ower meter correction applies to one port only; the other port is not
corrected.
Application and Operation Concepts 6-99

Alternate and Chop Sweep Modes
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Y ou can select the
ALTERNATE A and B or
CHOP A and B softkey within the Correction More
menu to activate either one or the other sweep modes . F or information about sweep types ,
refer to \Sweep Type Menu," located earlier in this chapter .
ALTERNATE A and B measures only one input per frequency sweep , in order to reduce
unwanted signals , such as crosstalk from sampler A to B when measuring B/R. Thus , this mode
optimizes the dynamic range for all four S-parameter measurements .
The disadvantages of this mode are associated with simultaneous transmission/re ection
measurements or full two-port calibrations: this mode takes twice as long as the chop mode to
make these measurements .
CHOP A and B is the default measurement mode . This mode measures both inputs A and B
during each sweep . Thus , if each channel is measuring a di erent parameter and both channels
are displayed, the chop mode o ers the fastest measurement time . This is the preferred
measurement mode for full two-port calibrations because both inputs remain active .
The disadvantage of this mode is that in measurements of high rejection devices greater than
85 dB , such as lters with a low-loss passband, maximum dynamic range may not be achieved.
Figure 6-56 shows the alternate sweep mode (bold trace) overlaying the chop sweep mode in
a band-pass lter measurement. Note the di erence in the crosstalk levels between the two
modes .

Figure 6-56. Alternate and Chop Sweeps Overlaid
6-100 Application and Operation Concepts
Alternate · Chop

Calibrating for Non-Insertable Devices
A test device having the same sex connector on both the input and output cannot be
connected directly into a transmission test con guration. Therefore , the device is considered
to be noninsertable, and one of the following calibration methods must be performed.
F or information on performing measurement calibrations , refer to Chapter 5, \Optimizing
Measurement Results ."
The adapter removal technique provides a means to accurately measure the noninsertable
device . F or each port, a separate 2-port error correction needs to be performed to create
a calibration set. The adapter removal algorithm uses the resultant data from the two
calibration sets and the nominal electrical length of the adapter to compute the adapters actual
S-parameters . This data is then used to generate a separate third cal set in which the forward
and reverse match and tracking terms are as if P ort 1 and P ort 2 could be connected. This is
possible because the actual S-parameters of the adapter are measured with great accuracy ,
thus allowing the e ects of the adapter to be completely removed when the third cal set is
generated. See Chapter 5.
With this method, you use two precision matched adapters which are \equal." T o be equal, the
adapters must have the same match, Z
, insertion loss , and electrical delay .
Modify the Cal Kit Thru De nition
With this method it is only necessary to use one adapter . The calibration kit thru de nition is
modi ed to compensate for the adapter and then saved as a user kit. However , the electrical
delay of the adapter must rst be found.
Application and Operation Concepts 6-101
A dapter Removal · Matched A dapters

Using the Instrument State Functions

Figure 6-57. Instrument State Function Block
The instrument state function block keys provide control of channel-independent system
functions . The following keys are described in this chapter:
: Limit lines and limit testing, time domain operation, and instrument modes .
: HP-IB controller modes , instrument addresses , and the use of the parallel port.
: T est sequencing.
Information on the remaining instrument state keys can be found in the following chapters:
6-102 Application and Operation Concepts

This section contains information on the following topics:
using the parallel port
This key is allows you to return the analyzer to local (front panel) operation from remote
(computer controlled) operation. This key will also abort a test sequence or hardcopy
print/plot. In this local mode , with a controller still connected on HP-IB , you can operate the
analyzer manually (locally) from the front panel. This is the only front panel key that is not
disabled when the analyzer is remotely controlled over HP-IB by a computer . The exception to
this is when local lockout is in e ect: this is a remote command that disables the
making it dicult to interfere with the analyzer while it is under computer control.
In addition, the
key provides access to the HP-IB menu, where you can set the
controller mode , and to the address menu, where you can enter the HP-IB addresses of
peripheral devices and select plotter/printer ports . Y ou can also set the mode of the parallel
port here .
The HP-IB menu consists of the following softkeys:
SYSTEM CONTROLLER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TALKER/LISTENER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
USE PASS CONTROL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SET ADDRESS
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PARALLEL [ ]
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
HP-IB DIAG on OFF
DISK UNIT NUMBER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
VOLUME NUMBER
The analyzer is factory-equipped with a remote programming interface using the
Hewlett-P ackard Interface Bus (HP-IB). This enables communication between the analyzer and
a controlling computer as well as other peripheral devices . This menu indicates the present
HP-IB controller mode of the analyzer . Three HP-IB modes are possible: system controller ,
talker/listener , and pass control.
Application and Operation Concepts 6-103

HP-IB ST A TUS Indicators
When the analyzer is connected to other instruments over HP-IB , the HP-IB ST A TUS indicators
in the instrument state function block light up to display the current status of the analyzer .
R = remote operation
L = listen mode
T = talk mode
S = service request (SRQ) asserted by the analyzer
SYSTEM CONTROLLER softkey activates the system controller mode . When in this mode ,
the analyzer can use HP-IB to control compatible peripherals , without the use of an external
computer . It can output measurement results directly to a compatible printer or plotter , store
instrument states using a compatible disk drive , or control a power meter for performing
service routines . The power meter calibration function requires system controller or pass
control mode .
TALKER/LISTENER softkey activates the talker/listener mode , which is the mode of
operation most often used. In this mode , a computer controller communicates with the
analyzer and other compatible peripherals over the bus . The computer sends commands or
instructions to and receives data from the analyzer . All of the capabilities available from the
analyzer front panel can be used in this remote operation mode , except for control of the
power line switch and some internal tests .
P ass Control Mode
USE PASS CONTROL softkey activates the third mode of HP-IB operation: the pass control
mode . In an automated system with a computer controller , the controller can pass control of
the bus to the analyzer on request from the analyzer . The analyzer is then the controller of the
peripherals , and can direct them to plot, print, or store without going through the computer .
When the peripheral operation is complete , control is passed back to the computer . Only one
controller can be active at a time . The computer remains the system controller , and can regain
control at any time .
Preset does not a ect the selected controller mode , but cycling the power returns the analyzer
to talker/listener mode .
Information on compatible peripherals is provided in Chapter 11, \Compatible P eripherals ."
This menu can be accessed by pressing the
SET ADDRESS softkey within the HP-IB menu.
In communications through the Hewlett-P ackard Interface Bus (HP-IB), each instrument on the
bus is identi ed by an HP-IB address . This decimal-based address code must be di erent for
each instrument on the bus .
This menu lets you set the HP-IB address of the analyzer , and enter the addresses of peripheral
devices so that the analyzer can communicate with them.
6-104 Application and Operation Concepts

Most of the HP-IB addresses are set at the factory and need not be modi ed for normal system
operation. The standard factory-set addresses for instruments that may be part of the system
are as follows:
Instrument HP-IB A ddress
External Disk Drive 00
P ower Meter 13
The address displayed in this menu for each peripheral device must match the address set on
the device itself. The analyzer does not have an HP-IB switch: its address is set only from the
front panel.
These addresses are stored in non-volatile memory and are not a ected by preset or by cycling
the power .
Using the P arallel P ort
The instrument's parallel port can be used in two di erent modes . By pressing
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNN
PARALLEL [ ] softkey , you can select either the
[COPY] mode or the
[GPIO] mode .
The copy mode allows the parallel port to be connected to a printer or plotter for the
outputting of test results . T o use the parallel port for printing or plotting, you must do the
following:
LOCAL
SET ADDRESSES .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PRINTER PORT .
PARALLEL so that copy is underlined.
The GPIO mode turns the parallel port into a \general purpose input/output" port.
In this mode the port can be connected to test xtures , power supplies , and other peripheral
equipment that might be used to interact with the analyzer during measurements . This mode is
exclusively used in test sequencing.
Application and Operation Concepts 6-105

key provides access to the system menu. This menu leads to additional menus
which control various aspects of the analyzer system. The following softkeys are located
within the system menu:
NNNNNNNNNNNNNNNNNNNNNNNNNNNN
SET CLOCK allows you to produce time stamps on plots and print-outs .
LIMIT MENU provides access to the limits menu.
TRANSFORM MENU (Option 010 Only) provides access to the transform menu.
HARMONIC MEAS (Option 002 Only) provides access to the harmonic mode menu.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
INSTRUMENT MODE provides access to the instrument mode menu.
SERVICE MENU provides access to the service menu (see the HP 8753D Network Analyzer
Service Guide).
This menu can be accessed by pressing
LIMIT MENU softkey within the system menu.
Y ou can have limit lines drawn on the display to represent upper and lower limits or device
speci cations with which to compare the test device . Limits are de ned in segments , where
each segment is a portion of the stimulus span. Each limit segment has an upper and a lower
starting limit value . Three types of segments are available: at line , sloping line , and single
point.
Limits can be de ned independently for the two channels , up to 22 segments for each channel.
These can be in any combination of the three limit types .
Limit testing compares the measured data with the de ned limits , and provides pass or fail
information for each measured data point. An out-of-limit test condition is indicated in ve
ways: with a FAIL message on the screen, with a beep , by changing the color of the failing
portions of a trace , with an asterisk in tabular listings of data, and with a bit in the HP-IB
event status register B . (The analyzer also has a BNC rear panel output that includes this status ,
but is only valid for a single channel measurement.)
The limit test output has three selectable modes . F or more information, refer to
Chapter 2, \Making Measurements ."
Limit lines and limit testing can be used simultaneously or independently . If limit lines are
on and limit testing is o , the limit lines are shown on the display for visual comparison and
adjustment of the measurement trace . However , no pass/fail information is provided. If limit
testing is on and limit lines are o , the speci ed limits are still valid and the pass/fail status is
indicated even though the limit lines are not shown on the display .
Limits are entered in tabular form. Limit lines and limit testing can be either on or o while
limits are de ned. As new limits are entered, the tabular columns on the display are updated,
and the limit lines (if on) are modi ed to the new de nitions . The complete limit set can be
o set in either stimulus or amplitude value .
Limits are checked only at the actual measured data points . It is possible for a device to be out
of speci cation without a limit test failure indication if the point density is insucient. Be sure
to specify a high enough number of measurement points in the stimulus menu.
6-106 Application and Operation Concepts
The System Menu · The · The Limits Menu

Limit lines are displayed only on Cartesian formats . In polar and Smith chart formats , limit
testing of one value is available: the value tested depends on the marker mode and is the
magnitude or the rst value in a complex pair . The message NO LIMIT LINES DISPLAYED is
shown on the display in polar and Smith chart formats .
The list values feature in the copy menu provides tabular listings to the display or a printer
for every measured stimulus value . These include limit line or limit test information if these
functions are activated. If limit testing is on, an asterisk is listed next to any measured value
that is out of limits . If limit lines are on, and other listed data allows sucient space , the upper
limit and lower limit are listed, together with the margin by which the device data passes or
fails the nearest limit.
If limit lines are on, they are plotted with the data on a plot. If limit testing is on, the PASS or
FAIL message is plotted, and the failing portions of the trace that are a di erent color on the
display are also a di erent color on the plot. If limits are speci ed, they are saved in memory
with an instrument state .
This menu allows you to specify limits for limit lines or limit testing, and presents a table of
limit values on the display . Limits are de ned in segments . Each segment is a portion of the
stimulus span. Up to 22 limit segments can be speci ed for each channel. The limit segments
do not have to be entered in any particular order: the analyzer automatically sorts them and
lists them on the display in increasing order of start stimulus value .
F or each segment, the table lists the segment number , the starting stimulus value , upper limit,
lower limit, and limit type . The ending stimulus value is the start value of the next segment,
or a segment can be terminated with a single point segment. Y ou can enter limit values as
upper and lower limits or delta limits and middle value . As new limit segments are de ned, the
tabular listing is updated. If limit lines are switched on, they are shown on the display .
If no limits have been de ned, the table of limit values shows the notation EMPTY . Limit
NNNNNNNNNNNNNN
segments are added to the table using the
ADD softkey or edited with the
EDIT softkey , as
previously described. The last segment on the list is followed by the notation END .
This menu sets the values of the individual limit segments . The segment to be modi ed, or a
default segment, is selected in the edit limits menu. The stimulus value can be set with the
controls in the entry block or with a marker (the marker is activated automatically when this
menu is presented). The limit values can be de ned as upper and lower limits , or delta limits
and middle value . Both an upper limit and a lower limit (or delta limits) must be de ned:
if only one limit is required for a particular measurement, force the other out of range (for
example +500 dB or 500 dB).
As new values are entered, the tabular listing of limit values is updated.
Segments do not have to be listed in any particular order: the analyzer sorts them
automatically in increasing order of start stimulus value when the
DONE key in the edit limits
menu is pressed. However , the easiest way to enter a set of limits is to start with the lowest
stimulus value and de ne the segments from left to right of the display , with limit lines turned
on as a visual check.
Phase limit values can be speci ed between +500
. Limit values above +180
are mapped into the range of 180
to correspond with the range of
phase data values .
Application and Operation Concepts 6-107

O set Limits Menu
This menu allows the complete limit set to be o set in either stimulus value or amplitude
value . This is useful for changing the limits to correspond with a change in the test setup , or
for device speci cations that di er in stimulus or amplitude . It can also be used to move the
limit lines away from the data trace temporarily for visual examination of trace detail.
6-108 Application and Operation Concepts

Knowing the Instrument Modes
There are ve major instrument modes of the analyzer:
frequency o set operation
harmonic mode operation (Option 002)
The instrument mode menu can be accessed by pressing
SYSTEM
INSTRUMENT MODE . This
menu contains the following softkeys:
NETWORK ANALYZER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
EXT SOURCE AUTO
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
EXT SOURCE MANUAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TUNED RECEIVER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FREQ OFFS MENU
This is the standard mode of operation for the analyzer , and is active after you press
or switch on the A C power .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SYSTEM
INSTRUMENT MODE
NETWORK ANALYZER returns the analyzer to the \normal"
network analyzer operating mode .
This mode uses the analyzer's internal source .
This mode allows the analyzer to phase lock to an external CW signal. External source mode is
best used for unknown signals , or for signals that drift. If a synthesized external source is used,
the tuned receiver mode is recommended because it is faster .
External source mode is useful in several applications:
when your test device is a mixer or other frequency translation device
in automated test applications where a source is already connected to the system, and you do
not want to switch between the system source and the analyzer's internal source .
Application and Operation Concepts 6-109

Typical T est Setup
Figure 6-58 shows a typical test setup using the external source mode . The same test setup is
applicable for either manual or automatic external source mode operation.

Figure 6-58. Typical Setup for the External Source Mode
External Source Mode In-Depth Description
Y ou may use the external source in automatic or manual mode . External source mode phase
locks the analyzer to an external CW signal.
The external source mode works only in CW time sweep .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
External Source A uto . If you press
INSTRUMENT MODE
EXT SOURCE AUTO the
analyzer turns on the external source auto mode . Y ou should observe the following points
when using this operation mode:
The auto mode has a wider capture range than the manual mode .
The manual mode is faster than the auto mode .
The auto mode searches for the incoming CW signal.
The capture range is typically 10% of the selected CW frequency .
This feature works only in CW time sweep type .
The incoming signal should not have large spurs or sidebands , as the analyzer may phase lock
on a spur or not phase lock at all.
The frequency the instrument has locked onto is shown on the analyzer , and is also available
via HP-IB .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
External Source Manual. If you press
INSTRUMENT MODE
EXT SOURCE MANUAL the
analyzer activates the external source manual mode . Y ou should observe the following points
when using this operation mode:
The manual mode has a smaller capture range than the auto mode .
The manual mode is much faster than auto mode .
This feature works only in CW time sweep type .
The incoming signal should not have large spurs or sidebands , as the analyzer may phase lock
on a spur or not phase lock at all.
6-110 Application and Operation Concepts
·

The frequency of the incoming signal should be within 0.5 to +5.0 MHz of the selected
frequency or the analyzer will not be able to phase lock to it.
CW Frequency Range in External Source Mode . 300 kHz to 3 GHz (6 GHz for Option 006)
Compatible Sweep Types . The external source mode will only function in CW time sweep .
If the instrument is in any other sweep type when external source is activated, the warning
message CHANGED TO CW TIME MODE will appear on the display .
External Source Requirements . The external source mode has spectral purity and power
input requirements , which are described in Chapter 7, \Speci cations and Measurement
Uncertainties ."
Capture Range . In either automatic or manual mode , you can enter the frequency of the
external CW signal using the
CW FREQ softkey (located under the Stimulus
key). The
actual signal must be within a certain frequency capture range as shown in T able 6-7.
T able 6-7. External Source Capture Ranges
Mode CW Frequency Capture Range
A utomatic 50 MHz 5 MHz of nominal CW frequency
> 50 MHz 10% of nominal CW frequency
Manual All 0.5 to +5 MHz of nominal CW frequency
If the incoming signal is not within the capture range , the analyzer will not phase lock
correctly .
Locking onto a signal with a frequency modulation component. Although the analyzer may
phase-lock onto a signal that has FM, it may not accurately show the signal's amplitude . The
accuracy of such measurements depends greatly on the IF bandwidth you choose . Use the
widest IF bandwidth available (3 kHz) if this problem occurs .
If you press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SYSTEM
INSTRUMENT MODE
TUNED RECEIVER the analyzer receiver operates
independently of any signal source .
The following features and limitations apply to the tuned receiver mode:
It is a fully synthesized receiver; it does not phase-lock to any source .
It functions in all sweep types .
It requires a synthesized CW source whose timebase is input to the analyzer's external
frequency reference .
F or more information on using the tuned receiver mode , refer to Chapter 2, \Making
Measurements ."
Application and Operation Concepts 6-111
Tuned Receiver Mode ·

Frequency O set Menu
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
If you press
INSTRUMENT MODE
FREQ OFFS MENU , the analyzer allows phase-locked
operation with a frequency o set between the internal source and receiver . This feature is
used in swept RF mixer measurements and has an upper frequency limit equal to that of the
analyzer being used.
This feature allows you to set the RF source to a xed o set frequency above or below the
receiver (as required in a mixer test, using a swept RF/IF and xed LO). Then you can input a
signal to a device over one frequency range and view its response over a di erent frequency
range . The maximum delay between the RF source and the R input is 0.3 microseconds . The
analyzer will show a signal that is a composite of the desired RF signal, image response , and
spurious signals .
Y ou can use the frequency o set in any sweep type in network analyzer mode . The two
user-de ned variables in this mode are receiver frequency and o set frequency (LO). The
analyzer automatically sets the source frequency equal to IF + LO or IF LO .
Mixer measurements and frequency o set mode applications are explained in application note
8753-2, \RF Component Measurements Mixer Measurements using the HP 8753B Network
Analyzer ," HP part number 5956-4362. This application note was written for the HP 8753B but
also applies to the HP 8753D . Also see product note 8753-2A, HP part number 5952-2771.
Frequency o set mode is useful for the following types of measurements on frequency-
translating device:
amplitude and phase tracking
Typical T est Setup
Figure 6-59 shows a typical test setup using frequency o set mode . Instructions are provided
in Chapter 3, \Making Mixer Measurements ." The attenuators shown reduce mismatch
uncertainties . The low pass lter keeps unwanted mixing products out of the sampler .
6-112 Application and Operation Concepts


Figure 6-59. Typical T est Setup for a Frequency O set Measurement
Frequency O set In-Depth Description
The source and receiver operate at two di erent frequencies in frequency o set operation.
The di erence between the source and receiver frequencies is the LO frequency that you
specify .
The two user-de ned variables in frequency o set are the receiver frequency , and the o set
(LO) frequency . The source frequency is automatically set by the instrument and equals
receiver frequency IF + LO or IF LO .
The Receiver Frequency . Y ou can choose a CW value or start and stop values for the receiver
frequency . The stimulus values , which appear on the analyzer display , will a ect only the
receiver .
The O set Frequency (LO). This frequency value is the di erence between the source and
receiver frequencies .
The analyzer's source locks to the receiver the LO frequency , regardless of
the o set value you selected.
Once the source is phase-locked and sweeping, the analyzer's source frequency
is not known precisely . As the LO frequency changes , the source tracks it to
maintain the receiver start/stop or CW frequency that you requested.
Frequency Hierarchy . The source frequency can be greater than or less than the LO
frequency . That is , the analyzer can measure either the lower or upper of the two IF mixing
products when it is in the frequency o set mode .
Frequency Ranges . Receiver frequency range: 300 kHz to 3 GHz (or 6 GHz with Option 006)
Compatible Instrument Modes and Sweep Types . Frequency o set is compatible with all
sweep types in the network analyzer mode .
Application and Operation Concepts 6-113

Receiver and Source Requirements . Refer to Chapter 7, \Speci cations and Measurement
Uncertainties ."
IF Input: R always; and port 1 or port 2 for a ratio measurement.
Display Annotations . The analyzer shows the annotation ofs when the frequency o set mode
is on. The annotation of? indicates that the source frequency is approximately 10 MHz away
from the sum of the IF and LO frequencies that you requested. This is most likely caused by
the LO frequency being outside the 1 to +5 MHz accuracy requirement.
Error Message . If you connect your test device before you switch on the frequency o set
function, the error message PHASE LOCK CAL FAILED appears on the screen. This is normal, and
will disappear when you press the
FREQ OFFS on OFF softkey .
Spurious Signal P assband Frequencies . Unwanted mixing products (or spurious LO signals)
at speci c frequencies can cause measurement inaccuracy , because of the characteristics of
a sampler . These speci c frequencies can be calculated. Y ou can reduce unwanted mixing
products going to the sampler by inserting a low pass lter at your test device's IF output.
6-114 Application and Operation Concepts

Harmonic Operation (Option 002 only)
The analyzer's harmonic menu can be accessed by pressing
HARMONIC MEAS .
The harmonic measurement mode allows you to measure the second or third harmonic as the
analyzer's source sweeps fundamental frequencies above 16 MHz. The analyzer can make
harmonic measurements in any sweep type .
Typical T est Setup
Figure 6-60. Typical Harmonic Mode T est Setup
Y ou can view the second or third harmonic alone by using only one of the analyzer's two
channels .
T o make the following types of measurements , uncouple channels 1 and 2, and switch on dual
channel.
The analyzer measures the fundamental on one channel while measuring the second or third
harmonic on the other channel.
The analyzer measures the second harmonic on one channel while measuring the third
harmonic on the other channel.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
COUPLE PWR ON off feature , the analyzer measures the fundamental on channel 1
while measuring the second or third harmonic in dBc on channel 2.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
COUPLE PWR ON off feature , the analyzer couples power between channels 1
and 2. This is useful when you are using the D2/D1 to D2 feature because you can change
fundamental power and see the resultant change in the harmonic power .
The analyzer shows the fundamental frequency value on the display . However , a marker in the
active entry area shows the harmonic frequency in addition to the fundamental. If you use
the harmonic mode , the annotation H=2 or H=3 appears on the left-hand side of the display .
The measured harmonic cannot not exceed the frequency limitations of the network analyzer's
receiver .
Coupling P ower Between Channels 1 and 2
COUPLE PWR ON off is intended to be used with the
D2/D1 toD2 on OFF softkey . Y ou
can use the D2/D1 to D2 function in harmonic measurements , where the analyzer shows
the fundamental on channel 1 and the harmonic on channel 2. D2/D1 to D2 ratios the two ,
Application and Operation Concepts 6-115

showing the fundamental and the relative power of the measured harmonic in dBc . Y ou must
uncouple channels 1 and 2 for this measurement, using the
COUPLED CHAN ON off softkey set
to OFF to allow alternating sweeps .
After uncoupling channels 1 and 2, you may want to change the fundamental power and see
the resultant change in relative harmonic power (in dBc).
COUPLE PWR ON off allows you to
change the power of both channels simultaneously , even though they are uncoupled in all other
respects .
The frequency range is determined by the upper frequency range of the instrument or system
(3 or 6 GHz) and by the harmonic being displayed. The 6 GHz operation requires an HP 8753D
Option 006. T able 6-8 shows the highest fundamental frequency for maximum frequency and
harmonic mode .
T able 6-8. Maximum Fundamental Frequency using Harmonic Mode
2nd Harmonic 1.5 GHz 3 GHz
3rd Harmonic 1.0 GHz 2.0 GHz
A ccuracy and input power
Refer to Chapter 7, \Speci cations and Measurement Uncertainties ." The maximum
recommended input power and maximum recommended source power are related
speci cations .
Using power levels greater than the recommended values , you may cause undesired harmonics
in the source and receiver . The recommended power levels ensure that these harmonics are
less than 45 dBc . Use test port power to limit the input power to your test device .
6-116 Application and Operation Concepts

Time Domain Operation (Option 010)
With Option 010, the analyzer can transform frequency domain data to the time domain or
time domain data to the frequency domain.
In normal operation, the analyzer measures the characteristics of a test device as a function
of frequency . Using a mathematical technique (the inverse F ourier transform), the analyzer
transforms frequency domain information into the time domain, with time as the horizontal
display axis . Response values (measured on the vertical axis) now appear separated in time
or distance , providing valuable insight into the behavior of the test device beyond simple
frequency characteristics .
An HP 8753D can be ordered with Option 010, or the option can be added at a
later date using the HP 85019B time domain retro t kit.
The transform used by the analyzer resembles time domain re ectometry (TDR) measurements .
TDR measurements , however , are made by launching an impulse or step into the test device
and observing the response in time with a receiver similar to an oscilloscope . In contrast, the
analyzer makes swept frequency response measurements , and mathematically transforms the
data into a TDR-like display .
The analyzer's transform menu can be accessed by pressing
SYSTEM
TRANSFORM MENU . This
menu consists of the following softkeys:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SET FREQ LOW PASS
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LOW PASS IMPULSE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LOW PASS STEP
NNNNNNNNNNNNNNNNNNNNNNNNNN
BANDPASS
NNNNNNNNNNNNNNNNNNNN
WINDOW
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SPECIFY GATE
The analyzer has three frequency-to-time transform modes:
Time domain bandpass mode is designed to measure band-limited devices and is the easiest
mode to use . This mode simulates the time domain response of an impulse input.
Time domain low pass step mode simulates the time domain response of a step input. As in
a traditional TDR measurement, the distance to the discontinuity in the test device , and the
type of discontinuity (resistive , capacitive , inductive) can be determined.
Time domain low pass impulse mode simulates the time domain response of an impulse
input (like the bandpass mode). Both low pass modes yield better time domain resolution for
a given frequency span than does the bandpass mode . In addition, when using the low pass
modes , you can determine the type of discontinuity . However , these modes have certain
limitations that are de ned in \Time domain low pass ," later in this section.
Application and Operation Concepts 6-117
The Transform Menu · · TRANSFORM ON off

The analyzer has one time-to-frequency transform mode:
F orward transform mode transforms CW signals measured over time into the frequency
domain, to measure the spectral content of a signal. This mode is known as the CW time
mode .
In addition to these transform modes , this section discusses special transform concepts such as
masking, windowing, and gating.
The relationship between the frequency domain response and the time domain response of
the analyzer is de ned by the F ourier transform. Because of this transform, it is possible to
measure , in the frequency domain, the response of a linear test device and mathematically
calculate the inverse F ourier transform of the data to nd the time domain response . The
analyzer's internal computer makes this calculation using the chirp-Z F ourier transform
technique . The resulting measurement is the fully error-corrected time domain re ection or
transmission response of the test device , displayed in near real-time .
Figure 6-61 illustrates the frequency and time domain re ection responses of a test device . The
frequency domain re ection measurement is the composite of all the signals re ected by the
discontinuities present in the test device over the measured frequency range .
In this section, all points of re ection are referred to as discontinuities .

Figure 6-61. Device Frequency Domain and Time Domain Re ection Responses
The time domain measurement shows the e ect of each discontinuity as a function of time
(or distance), and shows that the test device response consists of three separate impedance
changes . The second discontinuity has a re ection coecient magnitude of 0.035 (i.e . 3.5% of
the incident signal is re ected). Marker 1 on the time domain trace shows the elapsed time
from the reference plane (where the calibration standards are connected) to the discontinuity
that the signal travels at the speed of light. The signal travels slower than the speed of light
in most media (e .g. coax cables). This slower velocity (relative to light) can be compensated
for by adjusting the analyzer relative velocity factor . This procedure is described later in this
section under \Time domain bandpass ."
6-118 Application and Operation Concepts
General Theory

This mode is called bandpass because it works with band-limited devices . Traditional TDR
requires that the test device be able to operate down to dc . Using bandpass mode , there are no
restrictions on the measurement frequency range . Bandpass mode characterizes the test device
impulse response .
A djusting the Relative V elocity F actor
A marker provides both the time (x2) and the electrical length (x2) to a discontinuity . T o
determine the physical length, rather than the electrical length, change the velocity factor to
that of the medium under test:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MORE
VELOCITY FACTOR .
vacuum). Most cables have a velocity factor of 0.66 (polyethylene dielectrics) or 0.70 (te on
dielectrics).
T o cause the markers to read the actual one-way distance to a discontinuity ,
rather than the two-way distance , enter one-half the actual velocity factor .
Re ection Measurements Using Bandpass Mode
The bandpass mode can transform re ection measurements to the time domain. Figure 6-62a
shows a typical frequency response re ection measurement of two sections of cable .
Figure 6-62b shows the same two sections of cable in the time domain using the bandpass
mode .
Application and Operation Concepts 6-119
Time Domain Bandpass ·


Figure 6-62. A Re ection Measurement of Two Cables
The ripples in re ection coecient versus frequency in the frequency domain measurement are
caused by the re ections at each connector \beating" against each other .
One at a time , loosen the connectors at each end of the cable and observe the response in both
the frequency domain and the time domain. The frequency domain ripples increase as each
connector is loosened, corresponding to a larger re ection adding in and out of phase with
the other re ections . The time domain responses increase as you loosen the connector that
corresponds to each response .
Interpreting the bandpass re ection response horizontal axis . In bandpass re ection
measurements , the horizontal axis represents the time it takes for an impulse launched at the
test port to reach a discontinuity and return to the test port (the two-way travel time). In
Figure 6-61, each connector is a discontinuity .
Interpreting the bandpass re ection response vertical axis . The quantity displayed on the
vertical axis depends on the selected format. The common formats are listed in T able 6-9. The
default format is LOG MA G (logarithmic magnitude), which displays the return loss in decibels
(dB). LIN MA G (linear magnitude) is a format that displays the response as re ection coecient
(). This can be thought of as an average re ection coecient of the discontinuity over the
frequency range of the measurement. Use the REAL format only in low pass mode .
6-120 Application and Operation Concepts

T able 6-9. Time Domain Re ection F ormats
F ormat P arameter
LIN MA G Re ection Coecient (unitless) (0
REAL Re ection Coecient (unitless) (1
LOG MA G Return Loss (dB)
SWR Standing W ave Ratio (unitless)
Transmission Measurements Using Bandpass Mode
The bandpass mode can also transform transmission measurements to the time domain. F or
example , this mode can provide information about a surface acoustic wave (SA W) lter that
is not apparent in the frequency domain. Figure 6-63 illustrates a time domain bandpass
measurement of a 321 MHz SA W lter .

Figure 6-63. Transmission Measurement in Time Domain Bandpass Mode
Interpreting the bandpass transmission response horizontal axis . In time domain
transmission measurements , the horizontal axis is displayed in units of time . The time axis
indicates the propagation delay through the device . Note that in time domain transmission
measurements , the value displayed is the actual delay (not x2). The marker provides the
propagation delay in both time and distance .
Marker 2 in Figure 6-63 a indicates the main path response through the test device , which
has a propagation delay of 655.6 ns , or about 196.5 meters in electrical length. Marker 4 in
Figure 6-63b indicates the triple-travel path response at 1.91 s , or about 573.5 meters . The
response at marker 1 (at 0 seconds) is an RF feedthru leakage path. In addition to the triple
travel path response , there are several other multi-path responses through the test device ,
which are inherent in the design of a SA W lter .
Interpreting the bandpass transmission response vertical axis . In the log magnitude
format, the vertical axis displays the transmission loss or gain in dB; in the linear magnitude
format it displays the transmission coecient ( ). Think of this as an average of the
transmission response over the measurement frequency range .
Application and Operation Concepts 6-121

Time domain low pass
This mode is used to simulate a traditional time domain re ectometry (TDR) measurement. It
provides information to determine the type of discontinuity (resistive , capacitive , or inductive)
that is present. Low pass provides the best resolution for a given bandwidth in the frequency
domain. It may be used to give either the step or impulse response of the test device .
The low pass mode is less general-purpose than the bandpass mode because it places strict
limitations on the measurement frequency range . The low pass mode requires that the
frequency domain data points are harmonically related from dc to the stop frequency . That is ,
stop = n x start, where n = number of points . F or example , with a start frequency of 30 kHz
and 101 points , the stop frequency would be 3.03 MHz. Since the analyzer frequency range
starts at 30 kHz, the dc frequency response is extrapolated from the lower frequency data. The
requirement to pass dc is the same limitation that exists for traditional TDR.
Setting frequency range for time domain low pass
Before a low pass measurement is made , the measurement frequency range must meet the
(stop = n x start) requirement described above . The
SET FREQ LOW PASS softkey performs
the start frequency is set equal to stop/n.
If the low end of the measurement frequency range is critical, it is best to calculate
approximate values for the start and stop frequencies before pressing
SET FREQ LOW PASS
and calibrating. This avoids distortion of the measurement results . T o see an example ,
select the preset values of 201 points and a 300 kHz to 3 GHz frequency range . Now press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SET FREQ LOW PASS and observe the change in frequency values . The stop frequency changes
to 2.999 GHz, and the start frequency changes to 14.925 MHz. This would cause a distortion of
measurement results for frequencies from 300 kHz to 14.925 MHz.
If the start and stop frequencies do not conform to the low pass requirement
before a low pass mode (step or impulse) is selected and transform is turned
on, the analyzer resets the start and stop frequencies . If error correction is on
when the frequency range is changed, this turns it o .
T able 6-10. Minimum Frequency Ranges for Time Domain Low P ass
Number of P oints Minimum Frequency Range
3 30 kHz to 0.09 MHz
11 30 kHz to 0.33 MHz
26 30 kHz to 0.78 MHz
51 30 kHz to 1.53 MHz
101 30 kHz to 3.03 MHz
201 30 kHz to 6.03 MHz
401 30 kHz to 12.03 MHz
801 30 kHz to 24.03 MHz
1601 30 kHz to 48.03 MHz
6-122 Application and Operation Concepts

Minimum allowable stop frequencies . The lowest analyzer measurement frequency is 30
kHz, therefore for each value of n there is a minimum allowable stop frequency that can
be used. That is , the minimum stop frequency =n x 30 kHz. T able 6-10 lists the minimum
frequency range that can be used for each value of n for low pass time domain measurements .
Re ection Measurements In Time Domain Low P ass
Figure 6-64 shows the time domain response of an unterminated cable in both the low-pass step
and low-pass impulse modes .

Figure 6-64.
Time Domain Low P ass Measurements of an Unterminated Cable
Interpreting the low pass response horizontal axis . The low pass measurement horizontal
axis is the two-way travel time to the discontinuity (as in the bandpass mode). The marker
displays both the two-way time and the electrical length along the trace . T o determine the
actual physical length, enter the appropriate velocity factor as described earlier in this section
under \Time domain bandpass ."
Interpreting the low pass response vertical axis . The vertical axis depends on the chosen
format. In the low pass mode , the frequency domain data is taken at harmonically related
frequencies and extrapolated to dc . Because this results in the inverse F ourier transform
having only a real part (the imaginary part is zero), the most useful low pass step mode format
in this application is the real format. It displays the response in re ection coecient units .
This mode is similar to the traditional TDR response , which displays the re ected signal in a
real format (volts) versus time (or distance) on the horizontal axis .
The real format can also be used in the low pass impulse mode , but for the best dynamic range
for simultaneously viewing large and small discontinuities , use the log magnitude format.
F ault Location Measurements Using Low P ass
As described, the low pass mode can simulate the TDR response of the test device . This
response contains information useful in determining the type of discontinuity present.
Figure 6-65 illustrates the low pass responses of known discontinuities . Each circuit element
was simulated to show the corresponding low pass time domain S
response waveform.
The low pass mode gives the test device response either to a step or to an impulse stimulus .
Mathematically , the low pass impulse stimulus is the derivative of the step stimulus .
Application and Operation Concepts 6-123


Figure 6-65. Simulated Low P ass Step and Impulse Response W aveforms (Real F ormat)
Figure 6-66 shows example cables with discontinuities (faults) using the low pass step mode
with the real format.
6-124 Application and Operation Concepts


Figure 6-66. Low P ass Step Measurements of Common Cable F aults (Real F ormat)
Transmission Measurements In Time Domain Low P ass
Measuring small signal transient response using low pass step . Use the low pass mode to
analyze the test device's small signal transient response . The transmission response of a device
to a step input is often measured at lower frequencies , using a function generator (to provide
the step to the test device) and a sampling oscilloscope (to analyze the test device output
response). The low pass step mode extends the frequency range of this type of measurement to
3 GHz (6 GHz with an analyzer Option 006).
The step input shown in Figure 6-67 is the inverse F ourier transform of the frequency domain
response of a thru measured at calibration. The step rise time is proportional to the highest
frequency in the frequency domain sweep; the higher the frequency , the faster the rise time .
The frequency sweep in Figure 6-67 is from 10 MHz to 1 GHz.
Figure 6-67 also illustrates the time domain low pass response of an ampli er under test. The
average group delay over the measurement frequency range is the di erence in time between
the step and the ampli er response . This time domain response simulates an oscilloscope
measurement of the ampli er's small signal transient response . Note the ringing in the
ampli er response that indicates an under-damped design.
Application and Operation Concepts 6-125


Figure 6-67.
Time Domain Low P ass Measurement of an Ampli er Small Signal
Interpreting the low pass step transmission response horizontal axis . The low pass
transmission measurement horizontal axis displays the average transit time through the
test device over the frequency range used in the measurement. The response of the thru
connection used in the calibration is a step that reaches 50% unit height at approximately
time = 0. The rise time is determined by the highest frequency used in the frequency domain
measurement. The step is a unit high step , which indicates no loss for the thru calibration.
When a device is inserted, the time axis indicates the propagation delay or electrical length
of the device . The markers read the electrical delay in both time and distance . The distance
can be scaled by an appropriate velocity factor as described earlier in this section under \Time
domain bandpass ."
Interpreting the low pass step transmission response vertical axis . In the real format,
the vertical axis displays the transmission response in real units (for example , volts). F or the
ampli er example in Figure 6-67, if the ampli er input is a step of 1 volt, the output,
2.4 nanoseconds after the step (indicated by marker 1), is 5.84 volts .
In the log magnitude format, the ampli er gain is the steady state value displayed after the
initial transients die out.
Measuring separate transmission paths through the test device using low pass impulse
mode . The low pass impulse mode can be used to identify di erent transmission paths through
a test device that has a response at frequencies down to dc (or at least has a predictable
response , above the noise oor , below 30 kHz). F or example , use the low pass impulse mode to
measure the relative transmission times through a multi-path device such as a power divider .
Another example is to measure the pulse dispersion through a broadband transmission line ,
such as a ber optic cable . Both examples are illustrated in Figure 6-68 . The horizontal and
vertical axes can be interpreted as already described in this section under \Time Domain
Bandpass".
6-126 Application and Operation Concepts
Transient Response


Figure 6-68. Transmission Measurements Using Low P ass Impulse Mode
Masking occurs when a discontinuity (fault) closest to the reference plane a ects the response
of each subsequent discontinuity . This happens because the energy re ected from the rst
discontinuity never reaches subsequent discontinuities . F or example , if a transmission line has
two discontinuities that each re ect 50% of the incident voltage , the time domain response
(real format) shows the correct re ection coecient for the rst discontinuity (=0.50).
However , the second discontinuity appears as a 25% re ection (=0.25) because only half the
incident voltage reached the second discontinuity .
This example assumes a loss-less transmission line . Real transmission lines , with
non-zero loss , attenuate signals as a function of the distance from the reference
plane .
As an example of masking due to line loss , consider the time domain response of a 3 dB
attenuator and a short circuit. The impulse response (log magnitude format) of the short circuit
alone is a return loss of 0 dB , as shown in Figure 6-69a. When the short circuit is placed at
the end of the 3 dB attenuator , the return loss is 6 dB , as shown in Figure 6-69b . This value
actually represents the forward and return path loss through the attenuator , and illustrates
how a lossy network can a ect the responses that follow it.
Application and Operation Concepts 6-127
Time Domain Concepts · Masking


Figure 6-69. Masking Example
The analyzer provides a windowing feature that makes time domain measurements more useful
for isolating and identifying individual responses . Windowing is needed because of the abrupt
transitions in a frequency domain measurement at the start and stop frequencies . The band
limiting of a frequency domain response causes overshoot and ringing in the time domain
response , and causes a non-windowed impulse stimulus to have a sin(kt)/kt shape , where
k = /frequency span and t = time (see Figure 6-70). This has two e ects that limit the
usefulness of the time domain measurement:
Finite impulse width (or rise time). Finite impulse width limits the ability to resolve
between two closely spaced responses . The e ects of the nite impulse width cannot be
improved without increasing the frequency span of the measurement (see T able 6-11 ).

Figure 6-70. Impulse Width, Sidelobes , and Windowing
Sidelobes . The impulse sidelobes limit the dynamic range of the time domain measurement
by hiding low-level responses within the sidelobes of higher level responses . The e ects of
sidelobes can be improved by windowing (see T able 6-11).
Windowing improves the dynamic range of a time domain measurement by ltering the
frequency domain data prior to converting it to the time domain, producing an impulse
stimulus that has lower sidelobes . This makes it much easier to see time domain responses that
are very di erent in magnitude . The sidelobe reduction is achieved, however , at the expense of
6-128 Application and Operation Concepts
Windowing

increased impulse width. The e ect of windowing on the step stimulus (low pass mode only) is
a reduction of overshoot and ringing at the expense of increased rise time .
NNNNNNNNNNNNNNNNNNNN
T o select a window , press
TRANSFORM MENU
WINDOW . A menu is presented that allows
the selection of three window types (see T able 6-11 ).
T able 6-11. Impulse Width, Sidelobe Level, and Windowing V alues
Minimum 13 dB 0.60/Freq Span 21 dB 0.45/Freq Span
Normal 44 dB 0.98/Freq Span 60 dB 0.99/Freq Span
Maximum 75 dB 1.39/Freq Span 70 dB 1.48/Freq Span
impulse width. The bandpass impulse sidelobe levels are the same as low pass impulse sidelobe levels .
Choose one of the three window shapes listed in T able 6-11 . Or you can use the knob to select
any windowing pulse width (or rise time for a step stimulus) between the softkey values . The
time domain stimulus sidelobe levels depend only on the window selected.
NNNNNNNNNNNNNNNNNNNNNN
MINIMUM is essentially no window . Consequently , it gives the highest
sidelobes .
NORMAL (the preset mode) gives reduced sidelobes and is the mode
most often used.
MAXIMUM window gives the minimum sidelobes , providing the
greatest dynamic range .
USE MEMORY on OFF remembers a user-speci ed window pulse width (or step
rise time) di erent from the standard window values .
A window is activated only for viewing a time domain response , and does not a ect a
displayed frequency domain response . Figure 6-71 shows the typical e ects of windowing on
the time domain response of a short circuit re ection measurement.
Application and Operation Concepts 6-129


Figure 6-71. The E ects of Windowing on the Time Domain Responses of a Short Circuit
In the time domain, range is de ned as the length in time that a measurement can be made
without encountering a repetition of the response , called aliasing. A time domain response
repeats at regular intervals because the frequency domain data is taken at discrete frequency
points , rather than continuously over the frequency band.
M easur ement r ang e =
where F is the spacing between frequency data points
(number of points 1)
M easur ement r ang e =
f r eq uency span(H z )
example:
M easur ement = 201 points
1 M H z to 2: 001 GH z
(number of points 1)
R ang e =
f r eq uency span
= 100 10
E l ectr ical l eng th = r ang e the speed of l ig ht (3 10
= (100 10
s) (3 10
= 30 meter s
In this example , the range is 100 ns , or 30 meters electrical length. T o prevent the time domain
responses from overlapping, the test device must be 30 meters or less in electrical length for a
transmission measurement (15 meters for a re ection measurement). The analyzer limits the
stop time to prevent the display of aliased responses .
6-130 Application and Operation Concepts

T o increase the time domain measurement range , rst increase the number of points , but
remember that as the number of points increases , the sweep speed decreases . Decreasing the
frequency span also increases range , but reduces resolution.
Two di erent resolution terms are used in the time domain:
Response resolution. Time domain response resolution is de ned as the ability to resolve two
closely-spaced responses , or a measure of how close two responses can be to each other and
still be distinguished from each other . F or responses of equal amplitude , the response resolution
is equal to the 50% (6 dB) impulse width. It is inversely proportional to the measurement
frequency span, and is also a function of the window used in the transform. The approximate
formulas for calculating the 50% impulse width are given in T able 6-11 . F or example , using the
formula for the bandpass mode with a normal windowing function for a 50 MHz to 13.05 GHz
measurement (13.0 GHz span):
50% cal cul ated impul se w idth =
E l ectr ical l eng th (in air ) = (0: 151 10
s) (30 10
With this measurement, two equal responses can be distinguished when they are separated
by at least 4.53 centimeters . In a measurement with a 20 GHz span, two equal responses can
be distinguished when they are separated by at least 2.94 cm. Using the low pass mode (the
low pass frequencies are slightly di erent) with a minimum windowing function, you can
distinguish two equal responses that are about 1.38 centimeters or more apart.
F or re ection measurements , which measure the two-way time to the response , divide the
response resolution by 2. Using the example above , you can distinguish two faults of equal
magnitude provided they are 0.69 centimeters (electrical length) or more apart.
Remember , to determine the physical length, the relative velocity factor of
the transmission medium under test must be entered into the electrical length
equation.
F or example , a cable with a te on dielectric (0.7 relative velocity factor), measured under the
conditions stated above , has a fault location measurement response resolution of
0.45 centimeters . This is the maximum fault location response resolution. F actors such as
reduced frequency span, greater frequency domain data windowing, and a large discontinuity
shadowing the response of a smaller discontinuity , all act to degrade the e ective response
resolution.
Figure 6-72 illustrates the e ects of response resolution. The solid line shows the actual
re ection measurement of two approximately equal discontinuities (the input and output of an
SMA barrel). The dashed line shows the approximate e ect of each discontinuity , if they could
be measured separately .
Application and Operation Concepts 6-131


Figure 6-72. Response Resolution
While increasing the frequency span increases the response resolution, keep the following
points in mind:
The time domain response noise oor is directly related to the frequency domain data
noise oor . Because of this , if the frequency domain data points are taken at or below the
measurement noise oor , the time domain measurement noise oor is degraded.
The time domain measurement is an average of the response over the frequency range of
the measurement. If the frequency domain data is measured out-of-band, the time domain
measurement is also the out-of-band response .
Y ou may (with these limitations in mind) choose to use a frequency span that is wider than the
test device bandwidth to achieve better resolution.
Range resolution. Time domain range resolution is de ned as the ability to locate a single
response in time . If only one response is present, range resolution is a measure of how closely
you can pinpoint the peak of that response . The range resolution is equal to the digital
resolution of the display , which is the time domain span divided by the number of points on the
display . T o get the maximum range resolution, center the response on the display and reduce
the time domain span. The range resolution is always much ner than the response resolution
(see Figure 6-73).

Figure 6-73. Range Resolution of a Single Discontinuity
6-132 Application and Operation Concepts

Gating provides the exibility of selectively removing time domain responses . The remaining
time domain responses can then be transformed back to the frequency domain. F or re ection
(or fault location) measurements , use this feature to remove the e ects of unwanted
discontinuities in the time domain. Y ou can then view the frequency response of the remaining
discontinuities . In a transmission measurement, you can remove the e ects of multiple
transmission paths .
Figure 6-74a shows the frequency response of an electrical airline and termination.
Figure 6-74b shows the response in the time domain. The discontinuity on the left is due to the
input connector . The discontinuity on the right is due to the termination. W e want to remove
the e ect of the connector so that we can see the frequency response of just the airline and
termination. Figure 6-74c shows the gate applied to the connector discontinuity . Figure 6-74d
shows the frequency response of the airline and termination, with the connector \gated out."

Figure 6-74. Sequence of Steps in Gating Operation
Setting the gate . Think of a gate as a bandpass lter in the time domain (see Figure 6-75 ).
When the gate is on, responses outside the gate are mathematically removed from the time
domain trace . Enter the gate position as a start and stop time (not frequency) or as a center
and span time . The start and stop times are the bandpass lter 6 dB cuto times . Gates can
have a negative span, in which case the responses inside the gate are mathematically removed.
The gate's start and stop ags de ne the region where gating is on.
Application and Operation Concepts 6-133
Gating

Figure 6-75. Gate Shape
Selecting gate shape . The four gate shapes available are listed in T able 6-12 . Each gate has a
di erent passband atness , cuto rate , and sidelobe levels .
T able 6-12. Gate Characteristics
Gate Span Minimum 0.10 dB 48 dB 1.4/Freq Span 2.8/Freq Span
Normal 0.01 dB 68 dB 2.8/Freq Span 5.6/Freq Span
Wide 0.01 dB 57 dB 4.4/Freq Span 8.8/Freq Span
Maximum 0.01 dB 70 dB 12.7/Freq Span 25.4/Freq Span
The passband ripple and sidelobe levels are descriptive of the gate shape . The cuto time is the
time between the stop time (6 dB on the lter skirt) and the peak of the rst sidelobe , and
is equal on the left and right side skirts of the lter . Because the minimum gate span has no
passband, it is just twice the cuto time . Always choose a gate span wider than the minimum.
F or most applications , do not be concerned about the minimum gate span, simply use the knob
to position the gate markers around the desired portion of the time domain trace .
Transforming CW Time Measurements Into the Frequency Domain
The analyzer can display the amplitude and phase of CW signals versus time . F or example , use
this mode for measurements such as ampli er gain as a function of warm-up time (i.e . drift).
The analyzer can display the measured parameter (e .g. ampli er gain) for periods of up to
24 hours and then output the data to a digital plotter for hardcopy results .
These \strip chart" plots are actually measurements as a function of time (time is the
independent variable), and the horizontal display axis is scaled in time units . Transforms of
these measurements result in frequency domain data. Such transforms are called forward
transforms because the transform from time to frequency is a forward F ourier transform, and
can be used to measure the spectral content of a CW signal. F or example , when transformed
into the frequency domain, a pure CW signal measured over time appears as a single frequency
spike . The transform into the frequency domain yields a display that looks similar to a
spectrum analyzer display of signal amplitude versus frequency .
6-134 Application and Operation Concepts

F orward Transform Measurements
This is an example of a measurement using the F ourier transform in the forward direction,
from the time domain to the frequency domain (see Figure 6-76):

Figure 6-76. Ampli er Gain Measurement
Interpreting the forward transform vertical axis . With the log magnitude format selected,
the vertical axis displays dB . This format simulates a spectrum analyzer display of power versus
frequency .
Interpreting the forward transform horizontal axis . In a frequency domain transform of
a CW time measurement, the horizontal axis is measured in units of frequency . The center
frequency is the o set of the CW frequency . F or example , with a center frequency of 0 Hz, the
CW frequency (250 MHz in the example) is in the center of the display . If the center frequency
entered is a positive value , the CW frequency shifts to the right half of the display; a negative
value shifts it to the left half of the display . The span value entered with the transform on is
the total frequency span shown on the display . (Alternatively , the frequency display values can
be entered as start and stop .)
Demodulating the results of the forward transform. The forward transform can separate
the e ects of the CW frequency modulation amplitude and phase components . F or example , if
a test device modulates the transmission response (S
) with a 500 Hz AM signal, you can see
the e ects of that modulation as shown in Figure 6-77. T o simulate this e ect, apply a 500 Hz
sine wave to the analyzer rear panel EXT AM input.
Application and Operation Concepts 6-135


Figure 6-77. Combined E ects of Amplitude and Phase Modulation
Using the demodulation capabilities of the analyzer , it is possible to view the amplitude or
the phase component of the modulation separately . The window menu includes the following
softkeys to control the demodulation feature:
DEMOD: OFF
is the normal preset state , in which both the amplitude and phase components
of any test device modulation appear on the display .
AMPLITUDE
displays only the amplitude modulation, as illustrated in Figure 6-78a.
displays only the phase modulation, as shown in Figure 6-78b .

Figure 6-78.
Separating the Amplitude and Phase Components of T est-Device-Induced Modulation
F orward transform range . In the forward transform (from CW time to the frequency domain),
range is de ned as the frequency span that can be displayed before aliasing occurs , and is
similar to range as de ned for time domain measurements . In the range formula, substitute
time span for frequency span.
Example:
6-136 Application and Operation Concepts

N umber of points 1
R ang e =
= 1000 H er tz
F or the example given above , a 201 point CW time measurement made over a 200 ms time
span, choose a span of 1 kHz or less on either side of the center frequency (see Figure 6-79 ).
That is , choose a total span of 2 kHz or less .

Figure 6-79. Range of a F orward Transform Measurement
T o increase the frequency domain measurement range , increase the span. The maximum
range is inversely proportional to the sweep time , therefore it may be necessary to increase
the number of points or decrease the sweep time . Because increasing the number of points
increases the auto sweep time , the maximum range is 2 kHz on either side of the selected CW
time measurement center frequency (4 kHz total span). T o display a total frequency span of
4 kHz, enter the span as 4000 Hz.
Application and Operation Concepts 6-137

T est sequencing is an analyzer function that allows you to automate repetitive tasks . Y ou can
create a sequence as you are making a measurement. Then when you want to make that same
measurement again, you can recall the sequence and the analyzer will repeat the previous
keystrokes .
The following is a list of some of the key test sequencing features available on the
HP 8753D network analyzer:
Limited decision-making functions increase the versatility of the test sequences you create by
allowing you to jump from one sequence to another .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
GOSUB SEQUENCE function that allows you to call other sequences as sub-routines .
Y ou can create , title , save , and execute up to six sequences .
Y ou can save your sequences to a disk using the internal disk drive .
Y ou can use the parallel port as a general purpose input/output (GPIO) bus to read ve
TTL input bits in a decision making function, and send eight TTL output bits to control a
peripheral.
Product note 8753-3 \RF Component Measurements Applications of the
T est Sequence Function" provides practical applications examples for test
sequencing. This note was written for the HP 8753B but also applies to the
HP 8753D .
F eatures That Operate Di erently When Executed In a Sequence
The analyzer does not allow you to use the following keys in a sequence:
Commands That Sequencing Completes Before the Next Sequence Command
The analyzer completes all operations related to the following commands before continuing
with another sequence command:
recall or save (internal or external)
copy list values and operating parameters
CHAN1, CHAN2, W ait 0*
*W ait 0 is the special sequencing function
WAIT x with a zero entered for the delay value .
6-138 Application and Operation Concepts

Commands That Require a Clean Sweep
Many front panel commands disrupt the sweep in progress . F or example , changing the channel
or measurement type . When the analyzer does execute a disruptive command in a sequence ,
some instrument functions are inhibited until a complete sweep is taken. This applies to the
following functions:
F orward Stepping In Edit Mode
In the sequence modify mode , you can step through the selected sequence list, where the
analyzer executes each step .
A title may contain non-printable or special ASCII characters if you download it from an
external controller . A non-printable character is represented on the display as .
A sequence may contain up to 2 kbytes of instructions . Typically , this is around 200 sequence
command lines . T o estimate a sequence's size (in kbytes), use the following guidelines .
Type of Command Size in Bytes
Title string character 1
A ctive entry command 1 per digit
Embedding the V alue of the Loop Counter In a Title
Y ou can append a sequentially increasing or decreasing numeric value to the title of stored data
by placing a
MORE
NNNNNNNNNNNNNNNNN
TITLE
NNNNNNNNNNNNNN
MORE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LOOP COUNTER command after the title string. (Y ou
must limit the title to three characters if you will use it as a disk le name . The three-character
title and ve-digit loop counter number reach the eight-character limit for disk le names .)
This feature is useful in data logging applications .
Y ou can de ne a sequence to run automatically when you apply power to the analyzer . T o
make an autostarting sequence , create a sequence in position six and title it \A UTO". T o stop
an autostarting sequence , press
. T o stop an autostarting sequence from engaging at
power on, you must clear it from memory or rename it.
The instrument's parallel port can be used in two di erent modes . By pressing
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNN
PARALLEL [ ] softkey , you can select either the
[COPY] mode or the
[GPIO] mode .
The GPIO mode switches the parallel port into a \general purpose input/output" port.
In this mode , the port can be connected to test xtures , power supplies , and other peripheral
equipment that the analyzer can interact with through test sequencing.
Application and Operation Concepts 6-139

key accesses the Sequencing menu. This menu leads to a series of menus
that allow you to create and control sequences .
GOSUB SEQUENCE softkey , located in the Sequencing menu, activates a feature that allows
the sequence to branch o to another sequence , then return to the original sequence . F or
example , you could perform an ampli er measurement in the following manner:
the results . This sequence will act as a sub-routine .
measurements . In-between each power level setting, call sequence 1 as a sub-routine by
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
GOSUB SEQUENCE
SEQUENCE 1 . Now , sequence 2 will print the measurement
results for each input power level applied to the ampli er .
This menu can be accessed by pressing
TTL I/O in the Sequencing menu.
TTL Output for Controlling P eripherals
Eight TTL compatible output lines can be used for controlling equipment connected to the
parallel port. By pressing
NNNNNNNNNNNNNNNNNNNNNNN
TTL I/O you will access the softkeys (listed below) that control
the individual output bits . Refer to Figure 6-80 for output bus pin locations .
PARALLEL OUT ALL lets you input a number (0 to 255) in base 10 and outputs it to the bus
as binary .
SET BIT lets you set a single bit (0 - 7) to high on the output bus .
CLEAR BIT lets you set a single bit (0 - 7) to low on the output bus .
TTL Input Decision Making
Five TTL compatible input lines can be used for decision making in test sequencing. F or
example , if a test xture is connected to the parallel port and has a micro switch that needs to
be activated in order to proceed with a measurement, you can construct your test sequence
so that it checks the TTL state of the input line corresponding to the switch. Depending on
whether the line is high or low , you can jump to another sequence . T o access these decision
making functions , press
TTL I/O . Refer to Figure 6-80 for input bus pin locations .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PARALL IN BIT NUMBER lets you select the single bit (0 - 4) that the sequence will be
looking for .
PARALL IN IF BIT H lets you jump to another sequence if the single input bit you selected
is in a high state .
PARALL IN IF BIT L lets you jump to another sequence if the single input bit you selected
is in a low state .
Pin assignments:
pin 1 is the data strobe
pin 16 selects the printer
6-140 Application and Operation Concepts

pin 17 resets the printer
pins 18-25 are ground
Electrical speci cations for TTL high:
volts(H) = 2.7 volts (V)
current = 20 microamps (A)
Electrical speci cations for TTL low:
volts(L) = 0.4 volts (V)
current = 0.2 milliamps (mA)

Figure 6-80. P arallel P ort Input and Output Bus Pin Locations in GPIO Mode
Application and Operation Concepts 6-141

TTL OUT softkey provides access to the TTL out menu. This menu allows you to choose
between the following output parameters of the TTL output signal:
TTL OUT HIGH
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TTL OUT LOW
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
END SWEEP HIGH PULSE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
END SWEEP LOW PULSE
The TTL output signals are sent to the sequencing BNC rear panel output.
Sequencing Special Functions Menu
This menu is accessed by pressing the
SPECIAL FUNCTIONS softkey in the Sequencing menu.
Sequence Decision Making Menu
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
This menu is accessed by pressing the
DECISION MAKING softkey in the Sequencing Special
Functions menu.
Decision making functions are explained in more detail below . These functions check a
condition and jump to a speci ed sequence if the condition is true . The sequence called must
be in memory . A sequence call is a one-way jump . A sequence can jump to itself, or to any of
the other ve sequences currently in memory . Use of these features is explained under the
speci c softkey descriptions .
Decision making functions jump to a softkey location, not to a speci c sequence
Limit test, loop counter , and do sequence commands jump to any sequence residing in the
speci ed sequence position (1 through 6). These commands do not jump to a speci c sequence
title . Whatever sequence is in the selected softkey position will run when these commands are
executed.
Having a sequence jump to itself
A decision making command can jump to the sequence it is in. When this occurs , the sequence
starts over and all commands in the sequence are repeated. This is used a great deal in
conjunction with loop counter commands . See the loop counter description below .
TTL input decision making
TTL input from a peripheral connected to the parallel port (in the GPIO mode) can be used in a
decision making function. Refer to \The GPIO Mode " earlier in this section.
Limit test decision making
A sequence can jump to another sequence or start over depending on the result of a limit
test. When entered into a sequence , the
IF LIMIT TEST PASS and
IF LIMIT TEST FAIL
commands require you to enter the destination sequence .
6-142 Application and Operation Concepts

Loop counter decision making
The analyzer has a numeric register called a loop counter . The value of this register can be set
by a sequence , and it can be incremented or decremented each time a sequence repeats itself.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
The decision making commands
IF LOOP COUNTER = 0 and
IF LOOP COUNTER 0 jump
to another sequence if the stated condition is true . When entered into the sequence , these
commands require you to enter the destination sequence . Either command can jump to another
sequence , or restart the current sequence .
As explained earlier in \Embedding the V alue of The Loop Counter in The Title ," the loop
counter value can be appended to a title . This allows customized titles for data printouts or for
data les saved to disk.
Naming Files Generated by a Sequence
The analyzer can automatically increment the name of a le that is generated by a sequence
using a loop structure .
T o access the sequence lename menu, press:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FILE UTILITIES
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQUENCE FILENAMING
This menu presents two choices:
FILE NAME FILE0 supplies a name for the saved state and or data le . This also brings up
the Title File Menu.
PLOT NAME PLOTFILE supplies a name for the plot le generated by a plot-to-disk command.
This also brings up the Title File Menu.
The above keys show the current lename in the 2nd line of the softkey .
When titling a le for use in a loop function, you are restricted to only 2 characters in the
lename due to the 6 character length of the loop counter keyword \[LOOP]." When the le is
actually written, the [LOOP] keyword is expanded to only 5 ASCII characters (digits), resulting
in a 7 character lename .
After entering the 2 character lename , press:
NNNNNNNNNNNNNN
LOOP COUNTER
DONE
The analyzer allows you to use HP-GL (Hewlett-P ackard Graphics Language) to customize
messages or illustrations on the display of the analyzer . T o use HP-GL, the instrument must be
in system controller mode .
HP-GL commands should be entered into a title string using the
NNNNNNNNNNNNNNNNN
DISPLA Y
MORE
character selection menu.
Application and Operation Concepts 6-143

TITLE TO PERIPHERAL sequencing command (in the Sequencing Special Functions
menu) sends the HP-GL command string to the analyzer's HP-GL address . The address of the
analyzer's HP-GL graphics interface is always o set from the instrument's HP-IB address by 1:
If the current instrument address is an even number:
HP-GL address = instrument address +1.
If the current instrument address is an odd number:
HP-GL address = instrument address 1.
Two HP-GL commands require special consideration when used in local operation or in
sequencing. These are explained below:
The syntax for this command is P Ax,y where x and y are screen location coordinates
separated by a comma.
The syntax for this command is LB[text][etx]. The label command will print ASCII characters
until the etx command is seen. The etx is the ASCII value 3 (not the ASCII character 3).
The analyzer title function does not have the ASCII value 3, so the instrument allows the LB
END OF LABEL command (accessed by pressing
command to be terminated with the
NNNNNNNNNNNNNN
MORE
NNNNNNNNNNNNNNNNN
TITLE
NNNNNNNNNNNNNN
MORE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
END OF LABEL ).
Entering Sequences Using HP-IB
Y ou can create a sequence in a computer controller using HP-IB codes and enter it into the
analyzer over HP-IB . This method replaces the keystrokes with HP-IB commands . The following
is a procedure for entering a sequence over HP-IB:
Reading Sequences Using HP-IB
An external controller can read the commands in any sequence (in HP-IB command format).
Send the following command to the analyzer:
OUTPSEQx where x is a number from 1 to 6.
Allocate an adequate amount of string variable space in the external controller and execute an
ENTER statement.
6-144 Application and Operation Concepts
The · Special Commands ·

Ampli er T esting
The HP 8753D allows you to measure the transmission and re ection characteristics of many
ampli ers and active devices . Y ou can measure scalar parameters such as gain, gain atness ,
gain compression, reverse isolation, return loss (SWR), and gain drift versus time . A dditionally ,
you can measure vector parameters such as deviation from linear phase , group delay , complex
impedance and AM-to-PM conversion.

Figure 6-81. Ampli er P arameters
The analyzer allows you to make a swept-frequency measurement of an ampli er's second or
third harmonic as shown in Figure 6-82.

Figure 6-82.
Swept Frequency Ampli er Measurement of Absolute Fundamental, 2nd and 3rd
Application and Operation Concepts 6-145
Ampli er parameters · Harmonic Output Levels

The second/third harmonic response can be displayed directly in dBc , or dB below the
fundamental or carrier (see Figure 6-83). The ability to display harmonic level versus frequency
or RF power allows \real-time" tuning of harmonic distortion. Further , this swept harmonic
measurement, as well as all of the traditional linear ampli er measurements can be made
without reconnecting the test device to a di erent test con guration.

Figure 6-83.
Swept Frequency Ampli er Measurement of 2nd and 3rd Harmonic Distortion (dBc)
V ector network analyzers are commonly used to characterize ampli er gain compression versus
frequency and power level. This is essentially linear characterization since only the relative
level of the fundamental input to the fundamental output is measured. The narrowband
receiver is tuned to a precise frequency and, as a result, is immune from harmonic distortion.
Y ou may want to quantify the harmonic distortion itself.
Gain compression occurs when the input power of an ampli er is increased to a level that
reduces the gain of the ampli er and causes a nonlinear increase in output power . The point at
which the gain is reduced by 1 dB is called the 1 dB compression point. The gain compression
will vary with frequency , so it is necessary to nd the worst case point of gain compression in
the frequency band.
Once that point is identi ed, you can perform a power sweep of that CW frequency to measure
the input power at which the 1 dB compression occurs and the absolute power out (in dBm) at
compression.
6-146 Application and Operation Concepts
Gain Compression

Figure 6-84. Diagram of Gain Compression
Figure 6-85 illustrates a simultaneous measurement of fundamental gain compression and
second harmonic power as a function of input power .

Figure 6-85.
Swept P ower Measurement of Ampli er's Fundamental Gain Compression and 2nd
In a compression measurement it is necessary to know the RF input or output power at a
certain level of gain compression. Therefore , both gain and absolute power level need to be
accurately characterized. Uncertainty in a gain compression measurement is typically less
than 0.05 dB . Also , each input channel of the analyzer is calibrated to display absolute power
(typically within +0.5 dBm up to 3 GHz, and +1 dB up to 6 GHz).
Application and Operation Concepts 6-147
Harmonic Output Level

Metering the power level
When you are measuring a device that is very sensitive to absolute power level, it is important
that you accurately set the power level at either the device input or output. The analyzer
is capable of using an external HP-IB power meter and controlling source power directly .
Figure 6-86 shows a typical test con guration for setting a precise leveled input power at the
device input.

Figure 6-86.
T est Con guration for Setting RF Input using A utomatic P ower Meter Calibration
6-148 Application and Operation Concepts

Mixers or frequency converters , by de nition, exhibit the characteristic of having di erent
input and output frequencies . Mixer tests can be performed using the frequency o set
operation of the analyzer (with an external LO source) or using the tuned receiver operation of
the analyzer (with an external RF and LO source). The most common and convenient method
used is frequency o set.
F or a single-sideband mixer measurement, the RF source can be o set in frequency from the
input receiver frequency , allowing for a swept RF stimulus over one frequency range and
measurement of the IF response over another (in this case the output IF).
T o use the frequency o set guided setup for con guring a mixer measurement:
are calculated by the analyzer .)
The analyzer's tuned receiver mode allows you to tune its receiver to an arbitrary frequency
and measure signal power . This is only possible if the signal you want to analyze is at an exact
known frequency . Therefore , the RF and LO sources must be synthesized and synchronized
with the analyzer time base . Since the analyzer is not phaselocking in this con guration,
you can use it to measure conversion loss of a microwave mixer with an RF frequency range
output.
Y ou must take care to lter the output of the mixer because some of the
intermodulation and leakage products may be very close in frequency to the
desired IF . If these components are not ltered o , the analyzer may have
diculty selecting the correct signal to measure .
Tuned receiver mode also increases dynamic range . Broadband techniques like diode detection
have a high noise oor , while narrowband techniques like tuned receivers are much less
susceptible to noise .
Application and Operation Concepts 6-149
Mixer T esting · Frequency O set · Tuned Receiver

Mixer P arameters That Y ou Can Measure

Figure 6-87. Mixer P arameters
Transmission characteristics include conversion loss , conversion compression, group delay ,
and RF feedthru.
Re ection characteristics include return loss , SWR and complex impedance .
Characteristics of the signal at the output port include the output power , the spurious or
harmonic content of the signal, and intermodulation distortion.
Other parameters of concern are isolation terms , including LO to RF isolation and LO to IF
isolation.
In mixer transmission measurements , you have RF and LO inputs and an IF output. Also
emanating from the IF port are several other mixing products of the RF and LO signals . In
mixer re ection measurements , leakage signals from one mixer port propagate and appear
at the other two mixer ports . These unwanted mixing products or leakage signals can cause
distortion by mixing with a harmonic of the analyzer's rst down-conversion stage . T o ensure
that measurement accuracy is not degraded, you must lter certain frequencies or avoid them
by frequency selection. If you place attenuators at all mixer ports , you can reduce mismatch
uncertainties .
6-150 Application and Operation Concepts
A ccuracy Considerations

Attenuation at Mixer P orts
Mismatch between the instruments , cables , and mixer introduces errors in the measurement
that you cannot remove with a frequency response calibration. Y ou can reduce the mismatch
by using high quality attenuators as close to the mixer under test as possible .
When characterizing linear devices , you can use vector accuracy enhancement (measurement
calibration) to mathematically remove all systematic errors from the measurement, including
source and load mismatches . This is not possible when the device you are characterizing is a
mixer operating over multiple frequency ranges: therefore , source and load mismatches are not
corrected for and will add to overall measurement uncertainty .
T o reduce the measurement errors associated with the interaction between mixer port matches
and system port matches , you can place attenuators at all of the mixer's ports . Figure 6-88
shows a plot of swept conversion loss where no attenuation at mixer ports was used. The
ripple versus frequency is due to source and load mismatches .

Figure 6-88.
Conversion Loss versus Output Frequency Without Attenuators at
In contrast, Figure 6-90 made use of appropriate attenuation at all mixer ports . Y ou should give
extra care to the selection of the attenuator located at the mixer's IF port to avoid overdriving
the receiver . F or best results , choose the value of this attenuator so that the power incident on
the analyzer's R channel port is less than 10 dBm and greater than 35 dBm.
Application and Operation Concepts 6-151
Mixer P orts

Harmonics , linearity , and spurious signals also introduce errors that are not removed by
frequency response calibration. These errors are smaller with a narrowband detection scheme ,
but they may still interfere with your measurements . Y ou should lter the IF signal to reduce
these errors as much as possible .
Correct ltering between the mixer's IF port and the receiver's input port can eliminate
unwanted mixing and leakage signals from entering the analyzer's receiver . Figure 6-89 shows
a plot of mixer conversion loss when proper IF ltering was neglected.

Figure 6-89.
Example of Conversion Loss versus Output Frequency Without Correct
IF Signal P ath Filtering
Figure 6-90 shows the same mixer's conversion loss with the addition of a low pass lter at the
mixer's IF port.

Figure 6-90.
Example of Conversion Loss versus Output Frequency With Correct IF
Signal P ath Filtering and Attenuation at all Mixer P orts
Filtering is required in both xed and broadband measurements , but you can implement it more
easily in the xed situation. Therefore , when con guring broad-band (swept) measurements ,
you may need to trade some measurement bandwidth for the ability to more selectively lter
signals entering the analyzer's receiver .
6-152 Application and Operation Concepts
Filtering

By choosing test frequencies (frequency list mode), you can reduce the e ect of spurious
responses on measurements by avoiding frequencies that produce IF signal path distortion.
LO Frequency A ccuracy and Stability
The analyzer source is phaselocked to its receiver through a reference loop . In the frequency
o set mode , the mixer under test is inserted in this loop . T o ensure that the analyzer
phaselocks correctly , it is important that you use an LO source that has frequency accuracy
better than 1 MHz and residual FM < 20 kHz RMS.
Up-Conversion and Down-Conversion De nition
NNNNNNNNNNNNNNNNNNNNNNN
When you choose between
RF < LO and
RF > LO in the frequency o set menus , the analyzer
determines which direction the internal source must sweep in order to achieve the requested
IF frequency . F or example , to measure the lower sideband of a mixer , where the RF signal
is below the LO (RF < LO), the internal source must sweep backwards . See the examples in
Figure 6-91.

Figure 6-91. Examples of Up Converters and Down Converters
In standard mixer measurements , the input of the mixer is always connected to the analyzer's
RF source , and the output of the mixer always produces the IF frequencies that are received by
the analyzer's receiver .
Application and Operation Concepts 6-153
Frequency Selection

However , the ports labeled RF and IF on most mixers are not consistently connected to the
analyzer's source and receiver ports , respectively . These mixer ports are switched, depending
on whether a down converter or an upconverter measurement is being performed.
It is important to keep in mind that in the setup diagrams of the frequency o set mode , the
analyzer's source and receiver ports are labeled according to the mixer port that they are
connected to .
In a down converter measurement where the
DOWN CONVERTER softkey is selected, the
labeled RF , connecting to the mixer RF port, and the analyzer's receiver frequency is labeled
IF , connecting to the mixer IF port.
Because the RF frequency can be greater or less than the set LO frequency in this type of
NNNNNNNNNNNNNNNNNNNNNNN
measurement, you can select either
RF > LO or
RF < LO .
Figure 6-92. Down Converter P ort Connections
6-154 Application and Operation Concepts

In an up converter measurement where the
UP CONVERTER softkey is selected, the notation
on the setup diagram indicates that the analyzer's source frequency is labeled IF , connecting
to the mixer IF port, and the analyzer's receiver frequency is labeled RF , connecting to the
mixer RF port.
Because the RF frequency will always be greater than the set LO frequency in this type of
measurement, you must select only
RF > LO .
Figure 6-93. Up Converter P ort Connections
Application and Operation Concepts 6-155

Figure 6-94.
Example Spectrum of RF , LO , and IF Signals Present in a Conversion Loss Measurement
Conversion loss is a measure of how eciently a mixer converts energy from one frequency
to another . It is the ratio of the sideband output power to input signal power and is usually
expressed in dB .
Since the frequency response of the test system gets measured with the mixer's response , you
can perform a frequency response calibration to remove this group of errors .
Figure 6-95. Main Isolation T erms
Isolation is the amount of attenuation provided when a signal is applied to one port of a mixer
and measured at another port. Figure 6-95 shows the three main isolation terms .
LO F eedthru / LO to RF Leakage
LO feedthru, or LO-to-IF isolation, is the amount the LO signal that is attenuated when it
reaches the IF port.
LO to RF isolation is the amount the LO power is attenuated when it appears directly at the RF
port.
Both of these LO isolation terms are small for single and double balanced mixers . The RF signal
level applied to the mixer will have an a ect on this measurement. F or this reason, these
terms are usually measured with the RF port of the mixer terminated in a matched state .
6-156 Application and Operation Concepts
Conversion Loss · Isolation

RF feedthru, or RF-to-IF isolation, is the amount the RF power that is attenuated when it
reaches the IF port. This value is low in double balanced mixers . RF feedthru is usually less of
a problem than the LO isolation terms because the LO power level is signi cantly higher than
the RF power drive .
Y ou can make an RF feedthru measurement using the same instruments and setup that you use
to measure conversion loss . Because the source and receiver frequencies are the same , the
analyzer can use narrowband (tuned receiver) detection to make the measurement. The only
di erence that you need in the hardware con guration is that the IF lter needs to be removed
so the RF feedthru will not be ltered out.
The RF signal is applied to the RF port of the mixer and the feedthru is measured at the IF
port.
The RF feedthru level is very dependent on the LO signal that is applied. F or this reason, you
should make the measurement with the LO signal present at its normal operating level.
Y ou should perform a frequency response calibration to improve accuracy .
SWR / Return Loss
Re ection coecient () is de ned as the ratio between the re ected voltage (Vr) and incident
voltage (Vi). Standing wave ratio (SWR) is de ned as the ratio of maximum standing wave
voltage to the minimum standing wave voltage and can be derived from the re ection
coecient () using the equation shown below . Return loss can be derived from the re ection
coecient as well.
1 + j j
1 j j
R etur n l oss = 20 l og j j
Mixers are three-port devices , and the re ection from any one port
depends on the conditions of the other two ports . Y ou should replicate the
operating conditions the mixer will experience as closely as possible for the
measurement.
When you measure the RF port SWR, you should have the LO drive level present and set to the
expected frequency and power levels . The IF port should be terminated in a condition as close
to its operating state as possible .
The measurements of LO port SWR and IF port SWR are very similar . F or IF port SWR, you
should terminate the RF port in a matched condition and apply the LO signal at its normal
operating level. F or the LO port SWR, the RF and IF ports should both be terminated in
conditions similar to what will be present during normal operation.
Application and Operation Concepts 6-157
RF F eedthru · = · V r

Figure 6-96. Conversion Loss and Output P ower as a Function of Input P ower Level
Conversion compression is a measure of the maximum RF input signal level for which the
mixer will provide linear operation. The conversion loss is the ratio of the IF output level to
the RF input level, and this value remains constant over a speci ed input power range . When
the input power level exceeds a certain maximum, the constant ratio between IF and RF power
levels will begin to change . The point at which the ratio has decreased 1 dB is called the 1 dB
compression point.
level, until mixer compression begins and the mixer saturates .
Y ou can measure conversion compression using the same basic test con gurations that are used
to measure the conversion loss .
T o set up for a conversion compression measurement, rst measure the conversion loss of the
mixer under test. Set up for a CW measurement at the frequency of interest. Sweeping the
RF drive level over a 25 dB span soon shows the power level at which the conversion loss
increases by 1 dB .
With power meter calibration controlling the RF drive level, and the receiver calibrated to
measure output power , you can make accurate measurements of the output power at the 1 dB
compression point.
When you are making linear measurements , provide a reference for determining phase by
splitting the RF source power and send part of the signal into the reference channel. (This does
not work for frequency o set measurements , since the source and receiver are functioning at
di erent frequencies .)
T o provide a reference signal for the phase measurement, you need a second mixer . This mixer
is driven by the same RF and LO signals that are used to drive the mixer under test. The IF
output from the reference mixer is applied to the reference (R) channel of the analyzer .
6-158 Application and Operation Concepts
Conversion Compression · Phase Measurements

Amplitude and Phase Tracking
The match between mixers is de ned as the absolute di erence in amplitude and/or phase
response over a speci ed frequency range . The tracking between mixers is essentially how well
the devices are matched over a speci ed interval. This interval may be a frequency interval or
a temperature interval, or a combination of both.
Y ou can make tracking measurements by ratioing the responses of two mixer conversion loss
measurements . Then any di erence you view in response is due to the mixers and not the
measurement system.
Replace mixer A with the mixer that you want to compare it to . Mixer R should always remain
in place as the reference mixer .
Figure 6-97.
Connections for an Amplitude and Phase Tracking Measurement Between Two Mixers
Phase Linearity and Group Delay
Group delay is the rate of change of phase through a device with respect to frequency (d /d! ).
Traditionally , group delay has been used to describe the propagation delay ( g), and deviation
from linear phase through a linear device . However , this parameter also contains valuable
information about transmission delay and distortion through a non-linear device such as a
mixer or frequency converter . F or example , at group delay corresponds to low modulation
distortion (that is , carrier and sidebands propagate at the same rate).
Phase linearity and group delay are both measurements of the distortion of a transmitted
signal. Both measure the non-linearity of a device's phase response with respect to frequency .
Application and Operation Concepts 6-159

In standard vector error-correction, a thru (delay=0) is used as a calibration standard. The
solution to this problem is to use a calibration mixer with very small group delay as the
calibration standard.
An important characteristic to remember when selecting a calibration mixer is that the delay
of the device should be kept as low as possible . T o do this , select a mixer with very wide
bandwidth (wider bandwidth results in smaller delay).
6-160 Application and Operation Concepts

T o minimize the error introduced when you add an adapter to a measurement system, the
adapter needs to have low SWR or mismatch, low loss , and high repeatability .

Figure 6-98. A dapter Considerations
In a re ection measurement, the directivity of a system is a measure of the error introduced
by an imperfect signal separation device . It typically includes any signal that is detected at the
coupled port which has not been re ected by the test device . This directivity error will add
with the true re ected signal from the device , causing an error in the measured data. Overall
directivity is the limit to which a device's return loss or re ection can be measured. Therefore ,
it is important to have good directivity to measure low re ection devices .
F or example , a coupler has a 7 mm connector and 40 dB directivity , which is equivalent to a
re ection coecient of =0.01 (directivity in dB = 20 log ). Suppose we want to connect to
a device with an SMA male connector . W e need to adapt from 7 mm to SMA.
If we choose a precision 7 mm to SMA adapter with a SWR of 1.06, which has =0.03, the
overall directivity becomes =0.04 or 28 dB . However , if we use two adapters to do the same
job , the re ection from each adapter adds up to degrade the directivity to 17 dB . The last
example shown in Figure 6-98 uses three adapters that shows an even worse directivity of
14 dB . It is clear that a low SWR is desirable to avoid degrading the directivity of the system.
Application and Operation Concepts 6-161
Connection Considerations · A dapters

Fixtures are needed to interface non-coaxial devices to coaxial test instruments . It may also be
necessary to transform the characteristic impedance from standard 50 ohm instruments to a
non-standard impedance and to apply bias if an active device is being measured.
F or accurate measurements , the xture must introduce minimum change to the test signal, not
destroy the test device , and provide a repeatable connection to the device .
Hewlett-P ackard o ers several xtures for TO cans , stripline , and microstrip devices . Refer to
Chapter 11, \Compatible P eripherals ."
If Y ou W ant to Design Y our Own Fixture
Ideally , a xture should provide a transparent connection between the test instrument and
the test device . This means it should have no loss or electrical length and a at frequency
response , to prevent distortion of the actual signal. A perfect match to both the instrument
and the test device eliminates re ected test signals . The signal should be e ectively coupled
into the test device , rather than leaking around the device and resulting in crosstalk from input
to output. Repeatable connections are necessary to ensure consistent data.
Realistically , it is impossible to build an ideal xture , especially at high frequencies . However ,
it is possible to optimize the performance of the test xture relative to the performance of
the test device . If the xture's e ects on the test signal are relatively small compared to the
device's parameters , then the xture's e ects can be assumed to be negligible .
F or example , if the xture's loss is much less than the acceptable measurement uncertainty at
the test frequency , then it can be ignored.
6-162 Application and Operation Concepts
Fixtures

Hewlett-P ackard Company , \Simplify Y our Ampli er and Mixer T esting" 5956-4363
Hewlett-P ackard Company , \RF and Microwave Device T est for the '90s Seminar P apers"
Hewlett-P ackard Company \T esting Ampli ers and A ctive Devices with the HP 8720 Network
Analyzer" Product Note 8720-1 5091-1942E
Hewlett-P ackard Company \Mixer Measurements Using the HP 8753 Network Analyzer"
Product Note 8753-2A 5952-2771
General Measurement and Calibration T echniques
Rytting, Doug, \E ects of Uncorrected RF P erformance in a V ector Network Analyzer ," from
\Microwave Journal," April 1991
Blacka, Robert J ., \TDR Gated Measurements of Stripline T erminations ," Reprint from
\Microwave Product Digest," HP publication no . 5952-0359, March/April 1989
Montgomery , David, \Borrowing RF T echniques for Digital Design," Reprint from \Computer
Design" HP publication number 5952-9335, May 1982
Rytting, Doug, \A dvances in Microwave Error Correction T echniques ," Hewlett-P ackard RF and
Microwave Measurement Symposium paper HP publication number 5954-8378, June 1987
Rytting, Doug, \Improved RF Hardware and Calibration Methods for Network Analyzers ,"
Hewlett-P ackard RF and Microwave Measurement Symposium paper , 1991
Dunsmore , Joel, \A dd P ower-Meter A ccuracy to a Network Analyzer ," from \Microwaves and
RF ," January 1991
Fixtures and Non-Coaxial Measurements
Hewlett-P ackard Company , \Applying the HP 8510 TRL Calibration for Non-Coaxial
Measurements ," Product Note 8510-8A HP publication number 5091-3645E, F ebruary 1992
Hewlett-P ackard Company , \Measuring Chip Capacitors with the HP 8520C Network Analyzers
and Inter-Continental Microwave T est Fixtures" Product Note 8510-17 HP publication number
Hewlett-P ackard Company , \In-Fixture Microstrip Device Measurement Using TRL*
Calibration," Product Note 8720-2 HP publication number 5091-1943E, A ugust 1991
Hewlett-P ackard Company , \Calibration and Modeling Using the HP 83040 Modular Microcircuit
P ackage ," Product Note 83040-2 HP publication number 5952-1907, May 1990
\T est Fixtures and Calibration Standards ," Inter-Continental Microwave Product Catalog HP
Curran, Jim, \Network Analysis of Fixtured Devices ," Hewlett-P ackard RF and Microwave
Measurement Symposium paper , HP publication number 5954-8346, September 1986
Curran, Jim, \TRL Calibration for Non-Coaxial Measurements ," Hewlett-P ackard Semiconductor
T est Symposium paper
Elmore , Glenn and Louis Salz, \Quality Microwave Measurement of P ackaged A ctive Devices ,"
Hewlett-P ackard Journal, F ebruary 1987
\Measurement T echniques for Fixtured Devices ," HP 8510/8720 News HP publication number
Application and Operation Concepts 6-163

Hewlett-P ackard Company , \On-W afer Measurements Using the HP 8510 Network Analyzer and
Cascade Microtech W afer Probes ," Product Note 8510-6 HP publication number 5954-1579
Barr , J .T ., T . Burcham, A.C. Davidson, E.W . Strid, \A dvancements in On-W afer Probing
Calibration T echniques ," Hewlett-P ackard RF and Microwave Measurement Symposium paper ,
Lautzenhiser , S., A. Davidson, D . Jones , \Improve A ccuracy of On-W afer T ests Via LRM
Calibration," Reprinting from \Microwaves and RF" HP publication number 5952-1286, January
5091-6837, F ebruary 1993
6-164 Application and Operation Concepts
On-W afer Measurements

Speci cations and Measurement Uncertainties
The speci cations described in the table below apply to transmission measurements using 10 Hz
IF BW and full 2-port correction. Dynamic range is limited by the maximum test port power
and the receiver's noise oor .
T able 7-1. HP 8753D Dynamic Range
Frequency Range Dynamic Range
30 kHz to 300 kHz 100 dB
300 kHz to 1.3 GHz 110 dB
1.3 GHz to 3 GHz 110 dB
3 GHz to 6 GHz 105 dB
* 90 dB , 30 kHz to 50 kHz
y 100 dB , 300 kHz to 16 MHz, due to xed spurs
z 105 dB , Option 075
Speci cations and Measurement Uncertainties 7-1

HP 8753D Network Analyzer Speci cations
HP 8753D (50 ) with 7 mm T est P orts
The following speci cations describe the system performance of the HP 8753D network
analyzer . The system hardware includes the following:
Options: 006
Calibration kit: HP 85031B
Cables: HP 11857D
Measurement P ort Characteristics
The following tables describe the measurement port characteristics for both corrected and
uncorrected HP 8753D network analyzers .
T able 7-2.
Measurement P ort Characteristics (Corrected
) for HP 8753D (50 ) with
7 mm T est P orts
30 kHz to 300 kHz
300 kHz to 1.3 GHz 1.3 GHz to 3 GHz 3 GHz to 6 GHz
Directivity 55 dB 55 dB 51 dB 46 dB
Source match 55 dB 51 dB 49 dB 43 dB
Load match 55 dB 55 dB 51 dB 46 dB
Re ection tracking 0.001 dB 0.001 dB 0.005 dB 0.020 dB
Transmission tracking 0.008 dB 0.006 dB 0.009 dB 0.021 dB
* These characteristics apply for an environmental temperature of 25 5
C, with less than 1
the calibration temperature .
T able 7-3.
Measurement P ort Characteristics (Uncorrected
) for HP 8753D (50 )
with 7 mm T est P orts
30 kHz to 300 kHz
300 kHz to 1.3 GHz 1.3 GHz to 3 GHz 3 GHz to 6 GHz
35 dB 30 dB 25 dB
Source match 18 dB
16 dB 16 dB 14 dB
Load match 20 dB
18 dB 16 dB 14 dB
Re ection tracking 2.0 dB 1.5 dB 1.5 dB 2.5 dB
Transmission tracking 2.0 dB 1.5 dB 1.5 dB 2.5 dB
Crosstalk 100 dB 100 dB 100 dB 90 dB
Applies at 25 5
10 dB , 30 kHz to 50 kHz
15 dB , 30 kHz to 50 kHz
7-2 Speci cations and Measurement Uncertainties



Speci cations and Measurement Uncertainties 7-3
Transmission Measurement Uncertainties

Re ection Measurement Uncertainties


7-4 Speci cations and Measurement Uncertainties

HP 8753D (50 ) with Type-N T est P orts
The following speci cations describe the system performance of the HP 8753D network
analyzer . The system hardware includes the following:
Options: 006
Calibration kit: HP 85032B/E
Cables: HP 11851B
Measurement P ort Characteristics
The following table describes the measurement port characteristics for corrected HP 8753D
network analyzers .
T able 7-4.
Measurement P ort Characteristics (Corrected)
for HP 8753D (50 ) with
Type-N T est P orts
30 kHz to 300 kHz
300 kHz to 1.3 GHz 1.3 GHz to 3 GHz 3 GHz to 6 GHz
Directivity 50 dB 50 dB 47 dB 40 dB
Source match 49 dB 42 dB 36 dB 31 dB
Load match 50 dB 50 dB 47 dB 40 dB
Re ection tracking 0.005 dB 0.009 dB 0.019 dB 0.070 dB
Transmission tracking 0.014 dB 0.013 dB 0.026 dB 0.065 dB
Applies at 25 5
Speci cations and Measurement Uncertainties 7-5



7-6 Speci cations and Measurement Uncertainties
Transmission Measurement Uncertainties

Re ection Measurement Uncertainties


Speci cations and Measurement Uncertainties 7-7

HP 8753D (50 ) with 3.5 mm T est P orts
The following speci cations describe the system performance of the HP 8753D network
analyzer . The system hardware includes the following:
Options: 006
Calibration kit: HP 85033D
Cables: HP 11857D
Measurement P ort Characteristics
The following table describes the measurement port characteristics for corrected HP 8753D
network analyzers .
T able 7-5.
Measurement P ort Characteristics (Corrected)
for HP 8753D (50 ) with
3.5 mm T est P orts
30 kHz to 300 kHz
300 kHz to 1.3 GHz 1.3 GHz to 3 GHz 3 GHz to 6 GHz
Directivity 49 dB 46 dB 44 dB 38 dB
Source match 49 dB 44 dB 41 dB 37 dB
Load match 49 dB 46 dB 44 dB 38 dB
Re ection tracking 0.010 dB 0.005 dB 0.007 dB 0.009 dB
Transmission tracking 0.016 dB 0.014 dB 0.022 dB 0.048 dB
Applies at 25 5
7-8 Speci cations and Measurement Uncertainties



Speci cations and Measurement Uncertainties 7-9
Transmission Measurement Uncertainties

Re ection Measurement Uncertainties


7-10 Speci cations and Measurement Uncertainties

HP 8753D (75 ) with Type-N T est P orts
The following speci cations describe the system performance of the HP 8753D network
analyzer . The system hardware includes the following:
Options: 075
Calibration kit: HP 85036B
Cables: HP 11857B
Measurement P ort Characteristics
The following tables describe the measurement port characteristics for both corrected and
uncorrected HP 8753D network analyzers .
T able 7-6.
Measurement P ort Characteristics (Corrected)
for HP 8753D (75 ) with
Type-N T est P orts
30 kHz to 300 kHz
300 kHz to 1.3 GHz 1.3 GHz to 3 GHz
Directivity 48 dB 48 dB 43 dB
Source match 47 dB 41 dB 35 dB
Load match 48 dB 48 dB 43 dB
Re ection tracking 0.004 dB 0.010 dB 0.019 dB
Transmission tracking 0.018 dB 0.015 dB 0.033 dB
Applies at 25 5
T able 7-7.
Measurement P ort Characteristics (Uncorrected)
for HP 8753D (75 )
with Type-N T est P orts
30 kHz to 300 kHz
300 kHz to 1.3 GHz 1.3 GHz to 3 GHz
Directivity 34 dB 35 dB 30 dB
Source match 10 dB 16 dB 16 dB
Load match 14 dB 18 dB 16 dB
Re ection tracking 2.0 dB 1.5 dB 1.5 dB
Transmission tracking 2.0 dB 1.5 dB 1.5 dB
Crosstalk 100 dB 100 dB 100 dB
Applies at 25 5
Speci cations and Measurement Uncertainties 7-11



7-12 Speci cations and Measurement Uncertainties
Transmission Measurement Uncertainties

Re ection Measurement Uncertainties


Speci cations and Measurement Uncertainties 7-13

HP 8753D (75 ) with Type-F T est P orts
The following speci cations describe the system performance of the HP 8753D network
analyzer . The system hardware includes the following:
Options: 075
Calibration kit: HP 85039A
Cables: HP 11857B
Measurement P ort Characteristics
The following table describes the measurement port characteristics for corrected HP 8753D
network analyzers .
T able 7-8.
Measurement P ort Characteristics (Corrected)
for HP 8753D (75 )
using HP 85039A F-M T est P orts
30 kHz to 300 kHz
300 kHz to 1.3 GHz 1.3 GHz to 3 GHz
Directivity 38 dB 38 dB 32 dB
Source match 36 dB 36 dB 30 dB
Load match 38 dB 38 dB 32 dB
Re ection tracking 0.0080 dB 0.0080 dB 0.0320 dB
Transmission tracking 0.0618 dB 0.0346 dB 0.0778 dB
Applies at 25 5
7-14 Speci cations and Measurement Uncertainties



Speci cations and Measurement Uncertainties 7-15
Transmission Measurement Uncertainties

Re ection Measurement Uncertainties


7-16 Speci cations and Measurement Uncertainties

T able 7-9.
Measurement P ort Characteristics (Corrected)
for HP 8753D (75 )
using HP 85039A F-F T estports
30 kHz to 300 kHz
300 kHz to 1.3 GHz 1.3 GHz to 3 GHz
Directivity 38 dB 38 dB 32 dB
Source match 36 dB 36 dB 30 dB
Load match 32 dB 32 dB 26 dB
Re ection tracking 0.0080 dB 0.0080 dB 0.0320 dB
Transmission tracking 0.0959 dB 0.0518 dB 0.1118 dB
Applies at 25 5
Speci cations and Measurement Uncertainties 7-17



7-18 Speci cations and Measurement Uncertainties
Transmission Measurement Uncertainties

Re ection Measurement Uncertainties


Speci cations and Measurement Uncertainties 7-19

T able 7-10.
Measurement P ort Characteristics (Corrected)
for HP 8753D (75 )
using HP 85039A M-M T estports
30 kHz to 300 kHz
300 kHz to 1.3 GHz 1.3 GHz to 3 GHz
Directivity 32 dB 32 dB 26 dB
Source match 31.950 dB 31.170 dB 25.080 dB
Load match 38 dB 38 dB 32 dB
Re ection tracking 0.0348 dB 0.0253 dB 0.0728 dB
Transmission tracking 0.0780 dB 0.0474 dB 0.1106 dB
Applies at 25 5
7-20 Speci cations and Measurement Uncertainties



Speci cations and Measurement Uncertainties 7-21
Transmission Measurement Uncertainties

Re ection Measurement Uncertainties


7-22 Speci cations and Measurement Uncertainties

T able 7-11.
Measurement P ort Characteristics (Corrected)
for HP 8753D (75 )
using HP 85039A M-F T estports
30 kHz to 300 kHz
300 kHz to 1.3 GHz 1.3 GHz to 3 GHz
Directivity 32 dB 32 dB 26 dB
Source match 31.950 dB 31.170 dB 25.080 dB
Load match 32 dB 32 dB 26 dB
Re ection tracking 0.0348 dB 0.0253 dB 0.0728 dB
Transmission tracking 0.1121 dB 0.0646 dB 0.1445 dB
Applies at 25 5
Speci cations and Measurement Uncertainties 7-23



7-24 Speci cations and Measurement Uncertainties
Transmission Measurement Uncertainties

Re ection Measurement Uncertainties


Speci cations and Measurement Uncertainties 7-25

The speci cations listed in T able 1 range from those guaranteed by Hewlett-P ackard to those
typical of most HP 8753D instruments , but not guaranteed. Codes in the far right column
of T able 1 reference a speci cation de nition, listed below . These de nitions are intended
to clarify the extent to which Hewlett-P ackard supports the speci ed performance of the
HP 8753D .
service manual.
instrument cannot be veri ed outside the factory . Field procedures can verify performance
with a con dence prescribed by available standards .
tested speci cations , and can therefore be veri ed by functional pass/fail testing.
T: Typical but non-warranted performance characteristics intended to provide information
useful in applying the instrument. Typical characteristics are representative of most
instruments , though not necessarily tested in each unit. Not eld tested.
7-26 Speci cations and Measurement Uncertainties
Instrument Speci cations

T able 7-12. HP 8753D Instrument Speci cations (1 of 6)
Description Speci cation Code
Standard 30 kHz to 3 GHz S-1
Option 006 30 kHz to 6 GHz S-1
A ccuracy (at 25
C) 10 ppm S-1
C 7.5 ppm T
per year 3 ppm T
Resolution 1 Hz S-3
Range:
Standard 85 to +10 dBm S-1
Option 075 85 to +8 dBm S-1
Resolution 0.05 dB S-3
Level A ccuracy (at 0 dBm output level) (at 25
C)
1.0 dB S-1*
C)
15 to +5 dBm 0.2 dB (relative to 0 dBm output level) S-1
+5 to +10 dBm (Standard) 0.5 dB (relative to 0 dBm output level) S-1
+5 to +8 dBm (Option 075) 0.5 dB (relative to 0 dBm output level) S-1
>14 dB return loss to 6 GHz T
>16 dB return loss to 3 GHz T
2nd Harmonic (16 MHz to 3 GHz)
at +10 dBm output level
at 0 dBm output level
at 10 dBm output level
3rd Harmonic (16 MHz to 2 GHz)
at +10 dBm output level
at 0 dBm output level
at 10 dBm output level
at +10 dBm output level
at 10 dBm output level
* Explicitly tested as part of an on-site veri cation performed by Hewlett-P ackard.
Typical 30 kHz to 300 kHz and typical from 2 to 3 GHz for Option 075.
Speci cations and Measurement Uncertainties 7-27

T able 7-12. HP 8753D Instrument Speci cations (2 of 6)
Description Speci cation Code
Standard 30 kHz to 3 GHz S-1
Option 006 30 kHz to 6 GHz S-1
Impedance 50 ohms nominal (Standard)
Standard:
30 kHz to 50 kHz 10 dB return loss T
50 kHz to 300 kHz 18 dB return loss T
300 kHz to 1.3 GHz 18 dB return loss S-1
1.3 GHz to 3 GHz 16 dB return loss S-1
3 GHz to 6 GHz 14 dB return loss S-1
Impedance 75 ohms nominal (Option 075)
Option 075:
30 kHz to 300 kHz 10 dB return loss T
300 kHz to 3 GHz 16 dB return loss T
Maximum Input Level +10 dBm S-1
Damage Level +26 dBm or > 35 Vdc T
A verage Noise Level
300 kHz to 3 GHz
3 kHz IF bandwidth 82 dBm S-1*
10 Hz IF bandwidth 102 dBm S-1*
110 dBm T
3 GHz to 6 GHz
3 kHz IF bandwidth 77 dBm S-1*
10 Hz IF bandwidth 97 dBm S-1*
105 dBm T
Frequency Response (25 5
300 kHz to 3 GHz 1 dB S-1*
3 GHz to 6 GHz 2 dB S-1*
Internally Generated Harmonics (option 002)
at +8 dBm input level
at +0 dBm input level
at 15 dBm input level
at +8 dBm input level
at +0 dBm input level
at 15 dBm input level
Harmonic Measurement A ccuracy (25 5
16 MHz to 3 GHz 1 dB S-1
3 GHz to 6 GHz
3 dB S-1
Harmonic Measurement Dynamic Range
(with output at 10 dBm and input at < 15 dBm) 40 dBc T
* Explicitly tested as part of an on-site veri cation performed by Hewlett-P ackard.
Operation from 3 GHz to 6 GHz requires Option 006.
7-28 Speci cations and Measurement Uncertainties

T able 7-12. HP 8753D Instrument Speci cations (3 of 6)
Description Speci cation Code
Frequency O set Operation *
300 kHz to 6 GHz S-1
R Channel Input Requirements 0 to 35 dBm, to 3 GHz S-1
(required for phase-locked operation) 0 to 30 dBm, 3 GHz to 6 GHz S-1
0 to 30 dBm, 3 GHz to 6 GHz S-1
LO Spectral Purity and A ccuracy
Maximum Spurious Input
Residual FM
Frequency A ccuracy 1 to +1 MHz of nominal frequency T
A ccuracy (see Magnitude Characteristics and Phase Characteristics )
(CW Time sweep only)
300 kHz to 6 GHz S-1
P ower Level 0 to 25 dBm T
Maximum Spurious Input
Residual FM
A uto 500 ms T
Manual 50 ms T
Frequency Readout A ccuracy (auto) 0.1% T
Manual 0.5 to 5 MHz T
50 MHz 5 MHz of nominal CW frequency T
>50 MHz 10% of nominal CW frequency T
A ccuracy (see Magnitude Characteristics and Phase Characteristics )
* The HP 8753D RF source characteristics in this mode are dependent on the stability of the external LO source .
The RF source tracks the LO to maintain a stable IF signal at the R channel receiver input. Degradation in
accuracy is negligible with an HP 8642A/B or HP 8656B RF signal generator as the LO source .
Refer to \HP 8753D Descriptions and Options" for a functional description.
Operation from 3 GHz to 6 GHz requires option 006.
Measurement accuracy is dependent on the stability of the input signal.
Speci cations and Measurement Uncertainties 7-29

T able 7-12. HP 8753D Instrument Speci cations (4 of 6)
Description Speci cation Code
Display Resolution 0.01 dB/division S-3
Resolution 0.001 dB S-3
(10 Hz BW , inputs T est P ort 1 and T est P ort 2; R to 35 dBm) (see graph) S-1

Dynamic A ccuracy (Magnitude)
30 kHz to 3GHz
3 GHz to 6 GHz
Range 500 dB S-3
Resolution 0.001 dB S-3
30 kHz to 3 GHz 0.02 dB/
3 GHz to 6 GHz 0.04 dB/
PHASE CHARA CTERISTICS
S-3
Display Resolution 0.01
* Marker resolution for magnitude , phase , and delay is dependent upon the value measured; resolution is limited
to 5 digits .
CW sweep , +5 dBm into T est P ort, ratio measurement, 3 kHz BW
7-30 Speci cations and Measurement Uncertainties

T able 7-12. HP 8753D Instrument Speci cations (5 of 6)
INPUT GENERAL (cont.)
Description Speci cation Code
PHASE CHARA CTERISTICS (cont.)
(10 Hz BW , inputs T est P ort 1 and T est P ort 2; R to 35 dBm) (see graph) S-1

Dynamic A ccuracy (Phase)
Trace Noise (+5 dBm into T est P ort, ratio measurement)
30 kHz to 3 GHz
3 GHz to 6 GHz
Resolution 0.01
30 kHz to 3 GHz 0.05
3 GHz to 6 GHz 0.20
POLAR CHARA CTERISTICS (ratio measurement)
Range 10 10
up to 1000 units full scale S-3
Reference range of 500 units S-3
Speci cations and Measurement Uncertainties 7-31

T able 7-12. HP 8753D Instrument Speci cations (6 of 6)
INPUT GENERAL (cont.)
Description Speci cation Code
Group delay is computed by measuring the phase change within
a speci ed frequency step (determined by the frequency span
and the number of points per sweep).
Aperture (selectable) (frequency span)/(number of points 1) S-3
Maximum Aperture 20% of frequency span S-3
Range 1/2 (1/minimum aperture) S-3
(The maximum delay is limited to measuring no more than 180
of phase change within the minimum aperture .)
A ccuracy (see graph) S-3
The following graph shows group delay accuracy with an
HP 85047A T est Set with 7 mm full 2-port calibration and a 10
Hz IF bandwidth. Insertion loss is assumed to be
electrical length to be ten meters .

Group Delay A ccuracy vs . Aperture
In general, the following formula can be used to determine the accuracy , in seconds , of speci c group delay
measurement:
[0.003 Phase A ccuracy (deg)]/Aperture (Hz)
Depending on the aperture and device length, the phase accuracy used is either incremental phase accuracy or
worst case phase accuracy .
frequency span and the number of points per sweep).
Aperture (selectable) (frequency span)/(number of points 1) S-3
Maximum Aperture 20% of frequency span S-3
Range 1/2 (1/minimum aperture) S-3
(The maximum delay is limited to measuring no more than 180
of phase change within the minimum aperture .)
A ccuracy (see graph) S-3
7-32 Speci cations and Measurement Uncertainties
·

HP 8753D Network Analyzer General Characteristics
The following table shows typical measurement times for the HP 8753D network analyzer in
milliseconds .
T able 7-13. Typical Measurement Times
Typical Time for Completion (ms)
Number of P oints
51 201 401 1601
Uncorrected 125 200 300 900
125 200 300 900
245 510 855 2940
80 350 740 1790
Binary (Internal) 20 35 55 205
IEEE754 oating point format
32 bit 25 85 150 590
64 bit 40 115 220 840
S11 1-port calibration, with a 3 kHz IF bandwidth. Includes system retrace time , but does not include
bandswitch time . Time domain gating is assumed o .
S21 measurement with full 2-port calibration, using a 3 kHz IF bandwidth. Includes system retrace time and
RF switching time , but does not include bandswitch time . Time domain gating is assumed o .
Option 010 only , gating o .
Measured with HP 9000 series 300 computer .
HP-IB interface operates according to IEEE 488-1978 and IEC 625 standards and IEEE 728-1982
recommended practices .
Binary (internal 48-bit oating point complex format)
32/64 bit IEEE 754 Floating P oint F ormat
SH1, AH1, T6, TE0, L4, LE0, SR1, RL1, PP0, DC1, DT1, C1, C2, C3, C10, E2
Speci cations and Measurement Uncertainties 7-33

Front P anel Connectors
Connector Type : 7 mm precision
Impedance : 50 ohms (nominal)
Connector Conductor Depth : 0.000 to 0.003 in.
+15 V 2% 400 mA (combined load for both probe connections)
12.6 V 5.5% 300 mA (combined load for both probe connections)
Rear P anel Connectors
External Reference Frequency Input (EXT REF INPUT)
Frequency : 1, 2, 5, and 10 MHz (200 Hz at 10 MHz)
Level : 10 dBm to +20 dBm, typical
Impedance : 50 ohms
High-Stability Frequency Reference Output (10 MHz)(Option 001)
Frequency : 10.0000 MHz
C) : 0.05 ppm
Daily Aging Rate (after 30 days) : 3 10
Y early Aging Rate : 0.5 ppm/year
Output : 0 dBm minimum
Nominal Output Impedance : 50
External A uxiliary Input (A UX INPUT)
Input V oltage Limits : 10 V to +10 V
External AM Input (EXT AM)
1 volt into a 5 k resistor , 1 kHz maximum, resulting in approximately 8 dB/volt amplitude
modulation.
7-34 Speci cations and Measurement Uncertainties

External Trigger (EXT TRIGGER)
Triggers on a negative TTL transition or contact closure to ground.

Figure 7-1. External Trigger Circuit
T est Sequence Output (TEST SEQ)
This connector outputs a TTL signal which can be programmed by the user in a test sequence
to be high or low . By default, this output provides an end-of-sweep TTL signal. (F or use with
part handlers .)
Limit T est Output (LIMIT TEST)
This connector outputs a TTL signal of the limit test results . P ass: TTL high; F ail: TTL low .
T est P ort Bias Input (BIAS CONNECT)
Maximum voltage : +30 Vdc
Maximum current (no degradation in RF speci cations) : 200 mA
Maximum current : 1 A
Video Output (EXT MON)
The R, G, and B connectors drive external monitors with these characteristics:
R, G, B with synch on green.
75 ohm impedance .
1 Vp-p (0.7 V=white; 0 V=black; 0.3 V=synch).
This connector allows communication with compatible devices including external controllers ,
printers , plotters , disk drives , and power meters .
Speci cations and Measurement Uncertainties 7-35

P arallel P ort
This connector is used with parallel (or Centronics interface) peripherals such as printers
and plotters . It can also be used as a general purpose I/O port, with control provided by test
sequencing functions .
This connector is used with serial peripherals such as printers and plotters .
This connector is used for the optional A T compatible keyboard for titles and remote
front-panel operation.
48 to 66 Hz
115 V nominal (90 V to 132 V) or 230 V nominal (198 V to 264 V). 280 V A max.
Standards .
an anti-static bench mat (such as HP 92175T).
Dust: the environment should be as dust-free as possible .
Operating T emperature : 0
Error-Corrected T emperature Range : 1
C of calibration temperature
Humidity : 5% to 95% at 40
Altitude : 0 to 4500 meters (15,000 feet)
T emperature : 40
Humidity : 0 to 90% relative at +65
Altitude : 0 to 15,240 meters (50,000 feet)
7-36 Speci cations and Measurement Uncertainties

Net : 34 kg (75 lb)
Shipping : 37 kg (82 lb)
222 mm H 425 mm W 508 mm D
(8.75 16.75 20.0 in)
(These dimensions exclude front and rear panel protrusions .)

Data Retention Time with 3 V , 1.2 Ah Battery
T emperature at 70
C : 250 days (0.68 year) characteristically
T emperature at 40
C : 1244 days (3.4 years) characteristically
T emperature at 25
C : 10 years characteristically
Speci cations and Measurement Uncertainties 7-37



This chapter contains menus maps arranged in the following order:
CAL
COPY
DISPLA Y
F ORMA T
LOCAL
MARKER
MARKER F CTN
MEAS
MENU
SA VE/RECALL
PRESET
SCALE REF
SEQ
SYSTEM


d a · c b · 8-2 Menu Maps






d a · c b · Menu Maps 8-5





d a · c b · Menu Maps 8-7


d a · c b · 8-8 Menu Maps


d a · c b · Menu Maps 8-9


d a · c b · 8-10 Menu Maps


d a · c b · Menu Maps 8-11








8-14 Menu Maps


d a · c b · Menu Maps 8-15



K ey De nitions
This chapter contains information on the following topics:
Softkey and front-panel functions in alphabetical order (includes a brief description of each
Cross reference of programming commands to key functions
Cross reference of softkeys to front-panel access keys
SERVICE MENU keys are not included in this chapter . Service information
can be found in the HP 8753D Network Analyzer Service Guide .
Where to Look for More Information
A dditional information about many of the topics discussed in this chapter is located in the
following areas:
Chapter 2, \Making Measurements ," contains step-by-step procedures for making
measurements or using particular functions .
Chapter 4, \Printing, Plotting, and Saving Measurement Results ," contains instructions
for saving to disk or the analyzer internal memory , and printing and plotting displayed
measurements .
Chapter 5, \Optimizing Measurement Results ," describes techniques and functions for
achieving the best measurement results .
Chapter 6, \Application and Operation Concepts ," contains explanatory-style information
about many applications and analyzer operation.
HP 8753D Network Analyzer Programmer's Guide provides a complete description of all
HP-IB mnemonics .
K ey De nitions 9-1
function) · The

Guide T erms and Conventions
The eight keys along the right side of the analyzer display are called softkeys . Their labels
are shown on the display . The softkeys appear in shaded boxes in this chapter . F or example ,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TRANSMISSION . The labeled keys that are on the front panel of the analyzer are called
front-panel keys . The front-panel keys appear in unshaded boxes in this chapter . F or example ,
.
This section contains an alphabetical listing of softkey and front-panel functions , and a brief
description of each function.
is used to add a decimal point to the number you are entering.
is used to add a minus sign to the number you are entering.
is used to step up the current value of the active function.
The analyzer de nes the step for di erent functions . No units
terminator is required. F or editing a test sequence , this key can
be used to scroll through the displayed sequence .
is used to step down the current value of the active function.
The analyzer de nes the step for di erent functions . No units
terminator is required. F or editing a test sequence , this key can
be used to scroll through the displayed sequence .
is used to delete the last entry , or the last digit entered from
the numeric keypad. This key can also be used in two ways for
modifying a test sequence:
deleting a single-key command that you may have pressed by
mistake , (for example
deleting the last digit in a series of entered digits , as long as
you haven't yet pressed a terminator , (for example if you
2 but did not press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
goes to the delta marker menu, which is used to read the
di erence in values between the active marker and a reference
marker .
MODE OFF turns o the delta marker mode , so that the values displayed
for the active marker are absolute values .
REF = 1 establishes marker 1 as a reference . The active marker stimulus
and response values are then shown relative to this delta
reference . Once marker 1 has been selected as the delta
reference , the softkey label
REF = 1 is underlined in this
menu, and the marker menu is returned to the screen. In the
marker menu, the rst key is now labeled
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER REF = 1 .
The notation \REF=1" appears at the top right corner of the
graticule .
makes marker 2 the delta reference . A ctive marker stimulus
and response values are then shown relative to this reference .
9-2 K ey De nitions

REF = 3
makes marker 3 the delta reference .
makes marker 4 the delta reference .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
makes marker 5 the delta reference .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
REF = FIXED MKR
sets a user-speci ed xed reference marker . The stimulus and
response values of the reference can be set arbitrarily , and can
be anywhere in the display area. Unlike markers 1 to 5, the
xed marker need not be on the trace . The xed marker is
indicated by a small triangle , and the active marker stimulus
and response values are shown relative to this point. The
graticule .
Pressing this softkey turns on the xed marker . Its stimulus and
response values can then be changed using the xed marker
menu, which is accessed with the
softkey described below . Alternatively , the xed marker can be
set to the current active marker position, using the
MKR ZERO
softkey in the marker menu.
expresses the data in inverse S-parameter values , for use in
ampli er and oscillator design.
measures the absolute power amplitude at input A.
NNNNNNNNNNN
A/B
calculates and displays the complex ratio of input A to input B .
calculates and displays the complex ratio of the signal at input
A to the reference signal at input R.
puts the name of the active entry in the display title .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ACTIVE MRK MAGNITUDE
puts the active marker magnitude in the display title .
selects coaxial as the type of port used in adapter removal
calibration.
calibration.
ADAPTER DELAY is used to enter the value of electrical delay of the adapter used
in adapter removal calibration.
ADAPTER REMOVAL
provides access to the adapter removal menu.
displays the edit segment menu and adds a new segment to
the end of the list. The new segment is initially a duplicate of
the segment indicated by the pointer > and selected with the
SEGMENT softkey .
sets the HP-IB address of the analyzer , using the entry controls .
There is no physical address switch to set in the analyzer .
ADDRESS: CONTROLLER
sets the HP-IB address the analyzer will use to communicate
with the external controller .
with an external HP-IB disk drive .
K ey De nitions 9-3

with the power meter used in service routines .
presents a menu for adjusting display intensity , colors , and
accessing save and recall functions for modi ed LCD color sets .
leads to the beginning of the adjustment tests . These tests
generate correction constants that are used by the analyzer .
retrieves the full frequency list sweep .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ALTERNATE A and B
measures only one input per frequency sweep , in order to
reduce spurious signals . Thus , this mode optimizes the dynamic
range for all four S-parameter measurements .
AMPLITUDE OFFSET
adds or subtracts an o set in amplitude value . This allows
limits already de ned to be used for testing at a di erent
response level. F or example , if attenuation is added to or
removed from a test setup , the limits can be o set an equal
amount. Use the entry block controls to specify the o set.
ANALOG IN Aux Input
displays a dc or low frequency ac auxiliary voltage on the
vertical axis , using the real format. An external signal source
such as a detector or function generator can be connected to
the rear panel A UXILIARY INPUT connector .
ASSERT SRQ
sets the sequence bit in the Event Status Register , which can
be used to generate an SRQ (service request) to the system
controller .
AUTO FEED ON off turns the plotter auto feed function on or o when in the
de ne plot menu. It turns the printer auto feed on or o when
in the de ne print menu.
AUTO SCALE
brings the trace data in view on the display with one keystroke .
Stimulus values are not a ected, only scale and reference
values . The analyzer determines the smallest possible scale
factor that will put all displayed data onto 80% of the vertical
graticule . The reference value is chosen to put the trace in
center screen, then rounded to an integer multiple of the scale
factor .
AUX OUT on OFF allows you to monitor the analog bus nodes (except nodes
1,2,3,4,9,10,12) with external equipment. T o do this , connect
the equipment to the A UX INPUT BNC connector on the rear
panel.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
makes averaging factor the active function. Any value up to
999 can be used. The algorithm used for averaging is:
A(n) = S (n)=F + (1 1=F ) A(n 1)
A(n) = current average
S(n) = current measurement
F = average factor
turns the averaging function on or o for the active channel.
\A vg" is displayed in the status notations area at the left of the
display , together with the sweep count for the averaging factor ,
9-4 K ey De nitions
where · AVERAGING on OFF

when averaging is on. The sweep count for averaging is reset to
1 whenever an instrument state change a ecting the measured
data is made .
At the start of the averaging or following
averaging starts at 1 and averages each new sweep into the
trace until it reaches the speci ed averaging factor . The sweep
count is displayed in the status notations area below \A vg" and
updated every sweep as it increments . When the speci ed
averaging factor is reached, the trace data continues to be
updated, weighted by that averaging factor .
AVERAGING RESTART
averaging starts at 1 and averages each new sweep into the
trace until it reaches the speci ed averaging factor . The sweep
count is displayed in the status notations area below \A vg" and
updated every sweep as it increments .
is used to access three di erent noise reduction techniques:
sweep-to-sweep averaging, display smoothing, and variable IF
bandwidth. Any or all of these can be used simultaneously .
A veraging and smoothing can be set independently for each
channel, and the IF bandwidth can be set independently if the
stimulus is uncoupled.
B
measures the absolute power amplitude at input B .
calculates and displays the complex ratio of input B to input R.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
BACK SPACE
deletes the last character entered.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
sets the background intensity of the LCD as a percent of white .
The factory-set default value is stored in non-volatile memory .
BANDPASS
(Option 010 only) sets the time-domain bandpass mode .
toggles an annunciator which sounds to indicate completion of
BEEP DONE ON off
certain operations such as calibration or instrument state save .
turns the limit fail beeper on or o . When limit testing is on
BEEP FAIL on OFF
and the fail beeper is on, a beep is sounded each time a limit
test is performed and a failure detected. The limit fail beeper is
independent of the warning beeper and the operation complete
beeper .
toggles the warning annunciator . When the annunciator is on it
BEEP WARN on OFF
sounds a warning when a cautionary message is displayed.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
switches o the analyzer's display . This feature may be helpful
in prolonging the life of the LCD in applications where the
analyzer is left unattended (such as in an automated test
system). Pressing any front panel key will restore the default
display operation.
adjusts the brightness of the color being modi ed. See A djusting
Color for an explanation of using this softkey for color
modi cation of display attributes .
is used to enter the C0 term in the de nition of an OPEN
standard in a calibration kit, which is the constant term of the
cubic polynomial and is scaled by 10
K ey De nitions 9-5
· 15

is used to enter the C1 term, expressed in F/Hz (F arads/Hz) and
NNNNNNN
is used to enter the C2 term, expressed in F/Hz
C3 is used to enter the C3 term, expressed in F/Hz
key leads to a series of menus to perform measurement
calibrations for vector error correction (accuracy enhancement),
NNNNNNNNNN
and for specifying the calibration standards used. The
key also leads to softkeys which activate interpolated error
correction and power meter calibration.
CAL FACTOR
accepts a power sensor calibration factor % for the segment.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
brings up the segment modify menu and segment edit
(calibration factor menu) which allows you to enter a power
sensor's calibration factors . The calibration factor data entered
in this menu will be stored for power sensor A.
CAL FACTOR SENSOR B
brings up the segment modify menu and segment edit
(calibration factor menu) which allows you to enter a power
sensor's calibration factors . The calibration factor data entered
in this menu will be stored for power sensor B .
CAL KIT
leads to the select cal kit menu, which is used to select one of
the default calibration kits available for di erent connector
types . This , in turn, leads to additional menus used to de ne
calibration standards other than those in the default kits (refer
to Modifying Calibration Kits .) When a calibration kit has been
speci ed, its connector type is displayed in brackets in the
softkey label.
CAL KIT [ ]
selects the cal kit menu.
selects the HP 85056A/D cal kit.
selects the HP 85056K cal kit.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
selects the 2.92 mm cal kit model.
selects the HP 85033C cal kit.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
selects the HP 85033D cal kit.
selects the HP 85052C TRL cal kit.
selects the HP 85031B cal kit.
selects the HP 85032B cal kit.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
selects the HP 85036B/E cal kit.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL KIT: USER KIT
selects a kit other than those o ered by Hewlett-P ackard.
this default selection establishes the TRL/LRM LINE/MA TCH
standard as the characteristic impedance .
CAL ZO: SYSTEM ZO
allows you to modify the characteristic impedance of the
system for TRL/LRM calibration.
9-6 K ey De nitions

CALIBRATE MENU
leads to the calibration menu, which provides several accuracy
enhancement procedures ranging from a simple frequency
response calibration to a full two-port calibration. At the
completion of a calibration procedure , this menu is returned
to the screen, correction is automatically turned on, and the
CALIBRATE: NONE
is underlined if no calibration has been performed or if the
calibration data has been cleared. Unless a calibration is saved
in memory , the calibration data is lost on instrument preset,
power on, instrument state recall, or if stimulus values are
changed.
is used, along with the
key , to de ne the frequency range
of the stimulus . When the
key is pressed, its function
becomes the active function. The value is displayed in the
active entry area, and can be changed with the knob , step keys ,
or numeric keypad.
CENTER
sets the center frequency of a subsweep in a list frequency
sweep .
CHAN 1 allows you to select channel 1 as the active channel. The
active channel is indicated by an amber LED adjacent to the
corresponding channel key . The front panel keys allow you
to control the active channel, and all of the channel-speci c
functions you select apply to the active channel.
CH1 DATA [ ]
brings up the printer color selection menu. The channel 1 data
trace default color is magenta for color prints .
CH1 DATA LIMIT LN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
selects channel 1 data trace and limit line for display color
modi cation.
CH1 MEM selects channel 1 memory trace for display color modi cation.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
brings up the printer color selection menu. The channel 1
memory trace default color is green for color prints .
CHAN 2 allows you to select channel 2 as the active channel. The
active channel is indicated by an amber LED adjacent to the
corresponding channel key . The front panel keys allow you
to control the active channel, and all of the channel-speci c
functions you select apply to the active channel.
CH2 DATA [ ]
brings up the printer color selection menu. The channel 2 data
trace default color is blue for color prints .
CH2 DATA LIMIT LN
selects channel 2 data trace and limit line for display color
modi cation.
CH2 MEM [ ] brings up the printer color selection menu. The channel 2
memory trace default color is red for color prints .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CH2 MEM REF LINE
selects channel 2 memory and the reference line for display
color modi cation.
CH PWR [COUPLED] is used to apply the same power levels to each channel.
CH PWR [UNCOUPLED]
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
is used to apply di erent power levels to each channel.
K ey De nitions 9-7
· ·

CHOP A and B
measures A and B inputs simultaneously for faster
measurements .
CLEAR BIT when the parallel port is con gured for GPIO , 8 output bits can
be controlled with this key . When this key is pressed, \TTL
OUT BIT NUMBER" becomes the active function. This active
function must be entered through the keypad number keys ,
followed by the
key . The bit is cleared when the
is pressed. Entering numbers larger than 7 will result in bit 7
being cleared, and entering numbers lower than 0 will result in
bit 0 being cleared.
CLEAR LIST
deletes all segments in the list.
clears a sequence from memory . The titles of cleared sequences
will remain in load, store , and purge menus . This is done as a
convenience for those who often reuse the same titles .
COAX de nes the standard (and the o set) as coaxial. This causes the
analyzer to assume linear phase response in any o sets .
applies a linear phase compensation to the trace for use
with electrical delay . That is , the e ect is the same as if a
corresponding length of perfect vacuum dielectric coaxial
transmission line was added to the reference signal path.
provides access to the con gure menu. This menu contains
softkeys used to control raw o sets , spur avoidance , and the
test set transfer switch.
COLOR adjusts the degree of whiteness of the color being modi ed. See
\A djusting Color" for an explanation of using this softkey for
color modi cation of display attributes .
provides access to the con gure ext disk menu. This menu
contains softkeys used to the disk address , unit number , and
volume number .
CONTINUE SEQUENCE resumes a paused sequence .
located under the
key , is the standard sweep mode of
MENU
the analyzer , in which the sweep is triggered automatically and
continuously and the trace is updated with each sweep .
brings up the conversion menu which converts the measured
data to impedance (Z) or admittance (Y). When a conversion
parameter has been de ned, it is shown in brackets under the
softkey label. If no conversion has been de ned, the softkey
CONVERSION [OFF] .
provides access to the menus used for controlling external
plotters and printers and de ning the plot parameters .
turns error correction on or o . The analyzer uses the most
recent calibration data for the displayed parameter . If the
stimulus state has been changed since calibration, the original
state is recalled, and the message \SOURCE P ARAMETERS
CHANGED" is displayed.
COUNTER: ANALOG BUS
switches the counter to count the analog bus .
9-8 K ey De nitions

COUNTER: DIV FRAC N
switches the counter to count the A14 fractional-N V CO
frequency after it has been divided down to 100 kHz for
phase-locking the V CO .
COUNTER: FRAC N
switches the counter to count the A14 fractional-N V CO
frequency at the node shown on the overall block diagram.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
COUNTER: OFF
switches the internal counter o and removes the counter
display from the LCD .
COUPLED CH on OFF
toggles the channel coupling of stimulus values . With
COUPLED CH ON (the preset condition), both channels have
the same stimulus values (the inactive channel takes on the
stimulus values of the active channel).
is used to set the frequency for power sweep and CW time
sweep modes . If the instrument is not in either of these two
modes , it is automatically switched into CW time mode .
turns on a sweep mode similar to an oscilloscope . The analyzer
is set to a single frequency , and the data is displayed versus
time . The frequency of the CW time sweep is set with
CW FREQ
in the stimulus menu.
D2/D1 to D2 on OFF this math function ratios channels 1 and 2, and puts the results
in the channel 2 data array . Both channels must be on and
have the same number of points . Refer to Chapter 2, \Making
Measurements" for information on how to use this function to
make gain compression measurements .
displays both the current data and memory traces .
DATA ARRAY on OFF
speci es whether or not to store the error-corrected data on
disk with the instrument state .
DATA/MEM divides the data by the memory , normalizing the data to
the memory , and displays the result. This is useful for ratio
comparison of two traces , for instance in measurements of gain
or attenuation.
DATA - MEM subtracts the memory from the data. The vector subtraction is
performed on the complex data. This is appropriate for storing
a measured vector error , for example directivity , and later
subtracting it from the device measurement.
DATA ! MEMORY
stores the current active measurement data in the memory of
the active channel. It then becomes the memory trace , for use
in subsequent math manipulations or display . If a parameter
has just been changed and the * status notation is displayed at
the left of the display , the data is not stored in memory until
a clean sweep has been executed. The gating and smoothing
status of the trace are stored with the measurement data.
DATA ONLY on OFF stores only the measurement data of the device under test to a
disk le . The instrument state and calibration are not stored.
This is faster than storing with the instrument state , and uses
less disk space . It is intended for use in archiving data that will
later be used with an external controller , and cannot be read
back by the analyzer .
DECISION MAKING
presents the sequencing decision making menu.
K ey De nitions 9-9
DATA and MEMORY

DECR LOOP COUNTER
decrements the value of the loop counter by 1.
returns all the display color settings back to the factory-set
default values that are stored in non-volatile memory .
resets the plotting parameters to their default values .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DEFAULT PRNT SETUP
resets the printing parameters to their default values .
leads to the de ne save menu. Use this menu to specify the
data to be stored on disk in addition to the instrument state .
DEFINE PLOT leads to a sequence of three menus . The rst de nes which
elements are to be plotted and the auto feed state . The second
de nes which pen number is to be used with each of the
elements (these are channel dependent.) The third de nes the
line types (these are channel dependent), plot scale , and plot
speed.
leads to the de ne print menu. This menu de nes the printer
mode (monochrome or color) and the auto-feed state .
makes the standard number the active function, and brings up
the de ne standard menus . The standard number (1 to 8) is an
arbitrary reference number used to reference standards while
specifying a class .
selects the group delay format, with marker values given in
seconds .
DELAY/THRU de nes the standard type as a transmission line of speci ed
length, for calibrating transmission measurements .
Deletes the segment indicated by the pointer .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DELETE ALL FILES
deletes all les .
DELETE FILE
deletes a selected le .
sets the limits an equal amount above and below a speci ed
middle value , instead of setting upper and lower limits
separately . This is used in conjunction with
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER ! MIDDLE , to set limits for testing a device that is
speci ed at a particular value plus or minus an equal tolerance .
F or example , a device may be speci ed at 0 dB 3 dB . Enter
the delta limits as 3 dB and the middle value as 0 dB .
transform measurements .
DEMOD: OFF (Option 010 only) turns time domain demodulation o .
(Option 010 only) phase demodulation for CW TIME transform
measurements .
DIRECTORY SIZE lets you specify the number of directory les to be initialized
on a disk. This is particularly useful with a hard disk, where
you may want a directory larger than the default 256 les , or
with a oppy disk you may want to reduce the directory to
allow extra space for data les . The number of directory les
must be a multiple of 8. The minimum number is 8, and there
9-10 K ey De nitions
MIDDLE VALUE or

is no practical maximum limit. Set the directory size before
initializing a disk.
DISK UNIT NUMBER
speci es the number of the disk unit in the disk drive that is to
be accessed in an external disk store or load routine . This is
used in conjunction with the HP-IB address of the disk drive ,
and the volume number , to gain access to a speci c area on a
disk. The access hierarchy is HP-IB address , disk unit number ,
disk volume number .
DISP MKRS ON off displays response and stimulus values for all markers that are
turned on. A vailable only if no marker functions are on.
provides access to a series of menus for instrument and active
channel display functions . The rst menu de nes the displayed
active channel trace in terms of the mathematical relationship
between data and trace memory . Other functions include dual
channel display (overlaid or split), display intensity , color
selection, active channel display title , and frequency blanking.
DISPLAY: DATA
displays the current measurement data for the active channel.
leads to a series of service tests for the display .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DO BOTH FWD + REV
activates both forward and reverse calibration measurements
from selected calibration menus .
DO SEQUENCE has two functions:
It shows the current sequences in memory . T o run a
sequence , press the softkey next to the desired sequence title .
When entered into a sequence , this command performs a
one-way jump to the sequence residing in the speci ed
sequence position (SEQUENCE 1 through 6).
jumps to a softkey position, not to a speci c sequence title .
Whatever sequence is in the selected softkey position will
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DO SEQUENCE command is executed. This
command prompts the operator to select a destination
sequence position.
nishes one-port calibration (after all standards are measured)
and turns error correction on.
DONE 2-PORT CAL nishes two-port calibration (after all standards are measured)
and turns error correction on.
DONE RESP ISOL'N CAL nishes response and isolation calibration (after all standards
are measured) and turns error correction on.
nishes response calibration (after all standards are measured)
and turns error correction on.
DONE SEQ MODIFY terminates the sequencing edit mode .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
nishes TRL/LRM two-port calibration (after all standards are
measured) and turns error correction on.
sets the analyzer's source higher than the analyzer's receiver
for making measurements in frequency o set mode .
K ey De nitions 9-11
· run when the

DUAL CH on OFF
toggles between display of both measurement channels or
the active channel only . This is used in conjunction with
SPLIT DISP ON off in the display more menu to display both
channels . With
SPLIT DISP OFF the two traces are overlaid
on a single graticule .
duplicates a sequence currently in memory into a di erent
softkey position. Duplicating a sequence is straightforward.
F ollow the prompts on the analyzer screen. This command does
not a ect the original sequence .
EACH SWEEP
P ower meter calibration occurs on each sweep . Each
measurement point is measured by the power meter , which
provides the analyzer with the actual power reading. The
analyzer corrects the power level at that point. The number
of measurement/correction iterations performed on each
point is determined by the
NUMBER OF READINGS softkey .
This measurement mode sweeps slowly , especially when the
measured power is low . Low power levels require more time for
the power meter to settle . The power meter correction table in
memory is updated after each sweep . This table can be read or
changed via HP-IB .
displays a table of limit segments on the LCD , superimposed on
the trace . The edit limits menu is presented so that limits can
be de ned or changed. It is not necessary for limit lines or limit
testing to be on while limits are de ned.
presents the edit list menu. This is used in conjunction with
the edit subsweep menu to de ne or modify the frequency
sweep list. The list frequency sweep mode is selected with the
LIST FREQ softkey described below .
adjusts the electrical delay to balance the phase of the DUT . It
simulates a variable length lossless transmission line , which can
be added to or removed from a receiver input to compensate
for interconnecting cables , etc . This function is similar to
the mechanical or analog \line stretchers" of other network
analyzers . Delay is annotated in units of time with secondary
labeling in distance for the current velocity factor .
EMIT BEEP
NNNNNNNNNNNNNNNNNNNNNNNNNNNN
causes the instrument to beep once .
terminates the HP-GL \LB" command.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
END SWEEP HIGH PULSE
sets the TTL output on the test set interconnect to normally
high with a 10 s pulse high at the end of each sweep .
END SWEEP LOW PULSE
sets the TTL output on the test set interconnect to normally low
with a 10 s pulse low at the end of each sweep .
ERASE TITLE
deletes the entire title .
EXECUTE TEST
runs the selected service test.
selects the auto external source mode .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
selects the manual external source mode .
9-12 K ey De nitions

EXT TRIG ON POINT
is similar to the trigger on sweep , but triggers each data point in
a sweep .
EXT TRIG ON SWEEP
is used when the sweep is triggered on an externally generated
signal connected to the rear panel EXT TRIGGER input.
External trigger mode is allowed in every sweep mode .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
EXTENSION INPUT A
Use this feature to add electrical delay (in seconds) to extend
the reference plane at input A to the end of the cable . This is
used for any input measurements including S-parameters .
EXTENSION INPUT B
adds electrical delay to the input B reference plane for any B
input measurements including S-parameters .
EXTENSION PORT 1
extends the reference plane for measurements of S
S
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
EXTENSION PORT 2
extends the reference plane for measurements of S
S
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
toggles the reference plane extension mode . When this function
is on, all extensions de ned above are enabled; when o , none
of the extensions are enabled.
EXTERNAL DISK selects an (optional) external disk drive for SA VE/RECALL.
EXTERNAL TESTS
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
leads to a series of service tests .
appears during sequence modi cation, when external disk
is selected. FILE0 is the default name . A new name can be
entered when you save the state to disk.
FILE NAME FILE0
supplies a name for the saved istate and or data le . Brings up
the TITLE FILE MENU .
FILE UTILITIES provides access to the le utilities menu.
de nes the load in a calibration kit as a xed (not sliding) load.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FIXED MKR AUX VALUE
is used only with a polar or Smith format. It changes the
auxiliary response value of the xed marker . This is the second
part of a complex data pair , and applies to a magnitude/phase
marker , a real/imaginary marker , an R+jX marker , or a G+jB
marker . Fixed marker auxiliary response values are always
uncoupled in the two channels .
T o read absolute active marker auxiliary values following a
MKR ZERO operation, the auxiliary value can be reset to zero .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
leads to the xed marker menu, where the stimulus and
response values for a xed reference marker can be set
arbitrarily .
changes the stimulus value of the xed marker . Fixed marker
stimulus values can be di erent for the two channels if the
channel markers are uncoupled using the marker mode menu.
T o read absolute active marker stimulus values following a
MKR ZERO operation, the stimulus value can be reset to zero .
changes the response value of the xed marker . In a Cartesian
format this is the y-axis value . In a polar or Smith chart format
with a magnitude/phase marker , a real/imaginary marker , an
K ey De nitions 9-13
, and · , and · EXTENSIONS on OFF

R+jX marker , or a G+jB marker , this applies to the rst part of
the complex data pair . Fixed marker response values are always
uncoupled in the two channels .
T o read absolute active marker response values following a
MKR ZERO operation, the response value can be reset to zero .
FLAT LINE
de nes a at limit line segment whose value is constant with
frequency or other stimulus value . This line is continuous
to the next stimulus value , but is not joined to a segment
with a di erent limit value . If a at line segment is the nal
segment it terminates at the stop stimulus . A at line segment
is indicated as FL on the table of limits .
FORM FEED puts a form feed command into the display title .
presents a menu used to select the display format for the data.
V arious rectangular and polar formats are available for display
of magnitude , phase , impedance , group delay , real data, and
SWR.
FORMAT ARY on OFF speci es whether or not to store the formatted data on disk
with the instrument state .
FORMAT DISK brings up a menu for formatting a disk.
FORMAT: DOS
causes subsequent disk initialization to use the DOS disk format.
causes subsequent disk initialization to use the LIF disk format.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FORMAT: LIF is the default setting.
FORMAT EXT DISK
initializes media in external drive , and formats the disk using
the selected (DOS or LIF) format.
FORMAT INT DISK
initializes media in internal drive , and formats the disk using
the selected (DOS or LIF) format.
FORMAT INT MEMORY
clears all internal save registers and associated cal data and
memory traces .
FREQ OFFS MENU
leads to the frequency o set menu.
switches the frequency o set mode on and o .
FREQ OFFS on OFF
FREQUENCY
speci es the frequency of a calibration factor or loss value in
the power meter cal loss/sensor lists .
FREQUENCY BLANK
blanks the displayed frequency notation for security purposes .
Frequency labels cannot be restored except by instrument
preset or turning the power o and then on.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FREQUENCY: CW
sets the LO frequency to CW mode for frequency o set.
FREQUENCY: SWEEP
sets the LO frequency to sweep mode for frequency o set.
provides access to the series of menus used to perform a
complete calibration for measurement of all four S-parameters
of a two-port device . This is the most accurate calibration for
measurements of two-port devices .
measures the forward isolation of the calibration standard.
9-14 K ey De nitions

FWD MATCH (Label Class)
lets you enter a label for the forward match class . The label
appears during a calibration that uses this class .
FWD MATCH (Specify Class)
speci es which standards are in the forward match class in the
calibration kit.
FWD MATCH THRU is used to enter the standard numbers for the forward match
(thru) calibration. (F or default kits , this is the thru.)
FWD TRANS (Label Class)
lets you enter a label for the forward transmission class . The
label appears during a calibration that uses this class .
FWD TRANS (Specify Class)
speci es which standards are in the forward transmission class
in the calibration kit.
FWD TRANS THRU measures the forward frequency response in a two-port
calibration.
G+jB MKR displays the complex admittance values of the active marker in
rectangular form. The active marker values are displayed in
terms of conductance (in Siemens), susceptance , and equivalent
capacitance or inductance . Siemens are the international units
of admittance , and are equivalent to mhos (the inverse of
ohms). The Smith chart graticule is changed to admittance
form.
(Option 010 only) turns gating on or o in time domain mode .
(Option 010 only) allows you to specify the time at the center of
the gate .
GATE: SPAN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
(Option 010 only) allows you to specify the gate periods .
GATE: START
(Option 010 only) allows you to specify the starting time of the
gate .
GATE: STOP
(Option 010 only) allows you to specify the stopping time of the
gate .
(Option 010 only) leads to the gate shape menu.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
GATE SHAPE MAXIMUM
(Option 010 only) selects the widest time domain gate with the
smallest passband ripple .
GATE SHAPE MINIMUM
(Option 010 only) selects the narrowest time domain gate with
the largest passband ripple .
GATE SHAPE NORMAL
(Option 010 only) selects an intermediate time domain gate .
(Option 010 only) selects an intermediate time domain gate .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
GET SEQ TITLES
copies the sequence titles currently in memory into the six
softkey positions .
GOSUB SEQUENCE
calls sub-routines in sequencing.
speci es whether or not to store display graphics on disk with
the instrument state .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
GRATICULE [ ] brings up the color de nition menu. The graticule trace default
color is cyan.
K ey De nitions 9-15

GRATICULE TEXT
selects the graticule and a portion of softkey text (where there
is a choice of a feature being on or o ) for color modi cation.
F or example:
FREQUENCY BLANK on OFF .
HARMONIC MEAS
(Option 002 only) leads to the harmonics menu. This feature
phase locks to the 2nd or 3rd harmonic of the fundamental
signal. Measured harmonics cannot exceed the frequency range
of the analyzer receiver .
HARMONIC OFF
(Option 002 only) turns o the harmonic measurement mode .
(Option 002 only) selects measurement of the second harmonic .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
HARMONIC THIRD
(Option 002 only) selects measurement of the third harmonic .
provides an on-line quick reference guide to using the adapter
removal technique .
freezes the data trace on the display , and the analyzer stops
sweeping and taking data. The notation \Hld" is displayed at
the left of the graticule . If the * indicator is on at the left side
of the display , trigger a new sweep with
.
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
toggles the HP-IB diagnostic feature (debug mode). This mode
HP-IB DIAG on off
should only be used the rst time a program is written: if a
program has already been debugged, it is unnecessary .
When diagnostics are on, the analyzer scrolls a history of
incoming HP-IB commands across the display in the title line .
Nonprintable characters are represented as . If a syntax error
is received, the commands halt and a pointer ^ indicates the
misunderstood character . T o clear a syntax error , refer to the
\HP-IB Programming Reference" and \HP-IB Programming
Examples" chapters in the HP 8753D Network Analyzer
Programmer's Guide .
is used to select the bandwidth value for IF bandwidth
reduction. Allowed values (in Hz) are 3700, 3000, 1000, 300,
100, 30, and 10. Any other value will default to the closest
allowed value . A narrow bandwidth slows the sweep speed but
provides better signal-to-noise ratio . The selected bandwidth
value is shown in brackets in the softkey label.
jumps to one of the six sequence positions (SEQUENCE 1
through 6) if the limit test fails . This command executes any
sequence residing in the selected position. Sequences may
jump to themselves as well as to any of the other sequences in
memory . When this softkey is pressed, the analyzer presents
a softkey menu showing the six sequence positions and the
titles of the sequences located in them. Choose the destination
sequence to be called if the limit test fails .
jumps to one of the six sequence positions (SEQUENCE 1
through 6) if the limit test passes . This command executes any
sequence residing in the selected position. Sequences may
jump to themselves as well as to any of the other sequences in
memory . When this softkey is pressed, the analyzer presents a
softkey menu showing the six sequence positions , and the titles
9-16 K ey De nitions

of the sequences located in them. Choose the sequence to be
called if the limit test passes (destination sequence).
IF LOOP COUNTER = 0
prompts the user to select a destination sequence position
(SEQUENCE 1 through 6). When the value of the loop counter
reaches zero , the sequence in the speci ed position will run.
IF LOOP < > COUNTER 0
prompts the user to select a destination sequence position
(SEQUENCE 1 through 6). When the value of the loop counter
is no longer zero , the sequence in the speci ed position will run.
IMAGINARY
displays only the imaginary (reactive) portion of the measured
data on a Cartesian format. This format is similar to the real
format except that reactance data is displayed on the trace
instead of impedance data.
INCR LOOP COUNTER
increments the value of the loop counter by 1.
initializes the disk unit number and volume number selected in
the HP-IB menu, then returns to the disk menu. If more than
one hard disk volume is to be initialized, each volume must be
selected and initialized individually .
INITIALIZE DISK
leads to the initialize menu. Before data can be stored on
a disk, the disk must be initialized. If you attempt to store
without initializing the disk, the message \CA UTION: DISK
MEDIUM NOT INITIALIZED" is displayed. The disk format can
be selected to be either logical interchange format (LIF), or
DOS.
INPUT PORTS accesses a menu that allows you to measure the R, A, and B
channels .
INSTRUMENT MODE presents the instrument mode menu. This provides access to
the primary modes of operation (analyzer modes).
INTENSITY
sets the LCD intensity as a percent of the brightest setting. The
factory-set default value is stored in non-volatile memory .
INTERNAL TESTS
leads to a series of service tests .
selects the analyzer internal disk for the storage device .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
INTERNAL MEMORY
selects internal non-volatile memory as the storage medium for
subsequent save and recall activity .
turns interpolated error correction on or o . The interpolated
error correction feature allows the operator to calibrate the
system, then select a subset of the frequency range or a
di erent number of points . Interpolated error correction
functions in linear frequency , power sweep and CW time
modes . When using the analyzer in linear sweep , it is
recommended that the original calibration be performed with at
least 67 points per 1 GHz of frequency span.
leads to the isolation menu.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ISOLATION DONE
returns to the two-port cal menu.
measures the isolation of the device connected to the test port.
K ey De nitions 9-17
INTERPOL on OFF

ISTATE CONTENTS
describes the selected instrument state le (disk only)
translating the various lename pre xes into more descriptive
detail.
KIT DONE (MODIFIED)
terminates the cal kit modi cation process , after all standards
are de ned and all classes are speci ed. Be sure to save the kit
SAVE USER KIT softkey , if it is to be used later .
LABEL CLASS
leads to the label class menu, to give the class a meaningful
label for future reference during calibration.
LABEL CLASS DONE
nishes the label class function and returns to the modify cal
kit menu.
LABEL KIT leads to a menu for constructing a label for the user-modi ed
cal kit. If a label is supplied, it will appear as one of the ve
softkey choices in the select cal kit menu. The approach is
similar to de ning a display title , except that the kit label is
limited to ten characters .
LABEL STD The function is similar to de ning a display title , except that
the label is limited to ten characters .
LEFT LOWER draws a quarter-page plot in the lower left quadrant of the
page .
draws a quarter-page plot in the upper left quadrant of the
page .
leads to the o set limits menu, which is used to o set the
complete limit set in either stimulus or amplitude value .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
turns limit lines on or o . T o de ne limits , use the
LIMIT LINE on OFF
EDIT LIMIT LINE softkey described below . If limits have
been de ned and limit lines are turned on, the limit lines are
displayed on the LCD for visual comparison of the measured
data in all Cartesian formats .
If limit lines are on, they are plotted with the data on a plot,
and saved in memory with an instrument state . In a listing of
values from the copy menu with limit lines on, the upper limit
and lower limit are listed together with the pass or fail margin,
as long as other listed data allows sucient space .
LIMIT MENU
leads to a series of menus used to de ne limits or speci cations
with which to compare a test device . Refer to Limit Lines and
Limit T esting.
LIMIT TEST on OFF
turns limit testing on or o . When limit testing is on, the data
is compared with the de ned limits at each measured point.
Limit tests occur at the end of each sweep , whenever the data
is updated, when formatted data is changed, and when limit
testing is rst turned on.
Limit testing is available for both magnitude and phase values
in Cartesian formats . In polar and Smith chart formats , the
value tested depends on the marker mode and is the magnitude
or the rst value in a complex pair . The message \NO LIMIT
9-18 K ey De nitions

LINES DISPLA YED" is displayed in polar and Smith chart
formats if limit lines are turned on.
Five indications of pass or fail status are provided when limit
testing is on. A P ASS or F AIL message is displayed at the right
of the LCD . The trace vector leading to any measured point that
is out of limits is set to red at the end of every limit test, both
on a displayed plot and a hard copy plot. The limit fail beeper
sounds if it is turned on. In a listing of values using the copy
menu, an asterisk * is shown next to any measured point that is
out of limits . A bit is set in the HP-IB status byte .
LIMIT TEST RESULT
puts the result of a limit test into the display title .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
leads to the limit type menu, where one of three segment types
can be selected.
LIN FREQ activates a linear frequency sweep displayed on a standard
graticule with ten equal horizontal divisions . This is the default
preset sweep type .
LIN MAG
displays the linear magnitude format. This is a Cartesian format
used for unitless measurements such as re ection coecient
magnitude or transmission coecient magnitude , and for
linear measurement units . It is used for display of conversion
parameters and time domain transform data.
LIN MKR
displays a readout of the linear magnitude and the phase of the
active marker . Marker magnitude values are expressed in units;
phase is expressed in degrees .
LINE/MATCH
provides access to the Line/Match Menu for TRL/LRM
calibration.
LINE TYPE DATA selects the line type for the data trace plot. The default line
type is 7, which is a solid unbroken line .
LINE TYPE MEMORY
selects the line type for the memory trace plot. The default line
type is 7.
LIST
provides a tabular listing of all the measured data points and
their current values , together with limit information if it is
turned on. At the same time , the screen menu is presented, to
enable hard copy listings and access new pages of the table . 30
lines of data are listed on each page , and the number of pages is
determined by the number of measurement points speci ed in
the stimulus menu.
LIST FREQ provides a user-de nable arbitrary frequency list mode . This
list is de ned and modi ed using the edit list menu and the
edit subsweep menu. Up to 30 frequency subsweeps (called
\segments") of several di erent types can be speci ed, for
a maximum total of 1632 points . One list is common to both
channels . Once a frequency list has been de ned and a
measurement calibration performed on the full frequency list,
one or all of the frequency segments can be measured and
displayed without loss of calibration.
LIST VALUES
provides a tabular listing of all the measured data points and
their current values , together with limit information if it is
switched on. Thirty lines of data are listed on each page ,
K ey De nitions 9-19

and the number of pages is determined by the number of
measurement points speci ed in the stimulus menu.
LN/MATCH 1
measures the TRL/LRM line or match standard for PORT 1.
LN/MATCH 2
measures the TRL/LRM line or match standard for PORT 2.
turns the LO control mode on and o for frequency o set.
LO CONTROL on OFF
LO MENU
NNNNNNNNNNNNNNNNNNNNNN
leads to the LO menu. Allows you to con gure the external
source for frequency o set.
LO SOURCE ADDRESS
shows the HP-IB address of the LO source .
LOAD
de nes the standard type as a load (termination). Loads
are assigned a terminal impedance equal to the system
characteristic impedance Z0, but delay and loss o sets may still
be added. If the load impedance is not Z0, use the arbitrary
impedance standard de nition.
LOAD NO OFFSET
initiates measurement of a calibration standard load without
o set.
LOAD OFFSET initiates measurement of a calibration standard load with o set.
presents the load sequence from disk menu. Select the desired
sequence and the analyzer will load it from disk.
This key is used to return the analyzer to local (front panel)
operation from remote (computer controlled) operation. This
key will also abort a test sequence or hardcopy print/plot. In
this local mode , with a controller still connected on HP-IB , the
analyzer can be operated manually (locally) from the front
panel. This is the only front panel key that is not disabled
when the analyzer is remotely controlled over HP-IB by a
computer . The exception to this is when local lockout is in
making it dicult to interfere with the analyzer while it is
under computer control.
activates a logarithmic frequency sweep mode . The source is
stepped in logarithmic increments and the data is displayed on
a logarithmic graticule . This is slower than a continuous sweep
with the same number of points , and the entered sweep time
may therefore be changed automatically . F or frequency spans
of less than two octaves , the sweep type automatically reverts
to linear sweep .
displays the log magnitude format. This is the standard
Cartesian format used to display magnitude-only measurements
of insertion loss , return loss , or absolute power in dB versus
frequency .
displays the logarithmic magnitude value and the phase of the
active marker in P olar or Smith chart format. Magnitude values
are expressed in dB and phase in degrees . This is useful as a
fast method of obtaining a reading of the log magnitude value
without changing to log magnitude format.
displays the current value of the loop counter and allows you
to change the value of the loop counter . Enter any number
9-20 K ey De nitions
· key ,

from 0 to 32767 and terminate with the
key . The default
value of the counter is zero . This command should be placed in
a sequence that is separate from the measurement sequence .
decision command must call itself in order to function. The
LOOP COUNTER command must be in a separate sequence or the
counter value would always be reset to the initial value .
LOOP COUNTER
inserts the string \[LOOP]" into the lename .
NNNNNNNNNNNNN
accepts a power loss value for a segment in the power meter
cal power loss list. This value , for example , could be the
di erence (in dB) between the coupled arm and through arm of
a directional coupler .
LOSS/SENSR LISTS
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
presents the power loss/sensor lists menu. This menu performs
two functions:
Corrects coupled-arm power loss when a directional coupler is
used to sample the RF output.
Allows calibration factor data to be entered for one or two
power sensors .
Each function provides up to 12 separate frequency points ,
called segments , at which the user may enter a di erent power
loss or calibration factor . The instrument interpolates between
the selected points . Two power sensor lists are provided
because no single power sensor can cover the frequency range
possible with an HP 8753D .
LOW PASS IMPULSE
(Option 010 only) sets the transform to low pass impulse mode ,
which simulates the time domain response to an impulse input.
LOW PASS STEP
(Option 010 only) sets the transform to low pass step mode ,
which simulates the time domain response to a step input.
LOWER LIMIT
sets the lower limit value for the start of the segment in a
limit line list. If an upper limit is speci ed, a lower limit must
also be de ned. If no lower limit is required for a particular
measurement, force the lower limit value out of range (for
example 500 dB).
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MANUAL TRG ON POINT
waits for a manual trigger for each point. Subsequent pressing
of this softkey triggers each measurement. The annotation
\man" will appear at the left side of the display when the
instrument is waiting for the trigger to occur . This feature
is useful in a test sequence when an external device or
instrument requires changes at each point.
displays an active marker on the screen and provides access to
a series of menus to control from one to ve display markers
for each channel. Markers provide numerical readout of
measured values at any point of the trace .
The menus accessed from the
key provide several
basic marker operations . These include special marker modes
for di erent display formats , and a marker delta mode that
K ey De nitions 9-21

displays marker values relative to a speci ed value or another
marker .
MARKER ! AMP. OFS. uses the active marker to set the amplitude o set for the limit
lines . Move the marker to the desired middle value of the limits
and press this softkey . The limits are then moved so that they
are centered an equal amount above and below the marker at
that stimulus value .
MARKER ! CENTER changes the stimulus center value to the stimulus value of the
active marker , and centers the new span about that value .
MARKER ! CW
sets the CW frequency of the analyzer to the frequency
of the active marker . This feature is intended for use in
automated compression measurements . T est sequences allow
the instrument to automatically nd a maximum or minimum
point on a response trace . The
MARKER ! CW command sets
the instrument to the CW frequency of the active marker .
When power sweep in engaged, the CW frequency will already
be selected.
MARKER ! DELAY adjusts the electrical delay to balance the phase of the DUT .
This is performed automatically , regardless of the format and
the measurement being made . Enough line length is added to or
subtracted from the receiver input to compensate for the phase
slope at the active marker position. This e ectively attens
the phase trace around the active marker , and can be used
to measure electrical length or deviation from linear phase .
A dditional electrical delay adjustments are required on DUT s
without constant group delay over the measured frequency
span. Since this feature adds phase to a variation in phase
versus frequency , it is applicable only for ratioed inputs .
MARKER ! MIDDLE sets the midpoint for
DELTA LIMITS using the active marker
to set the middle amplitude value of a limit segment. Move the
marker to the desired value or device speci cation, and press
this key to make that value the midpoint of the delta limits .
The limits are automatically set an equal amount above and
below the marker .
MARKER ! REFERENCE makes the reference value equal to the active marker's
response value , without changing the reference position. In a
polar or Smith chart format, the full scale value at the outer
circle is changed to the active marker response value . This
softkey also appears in the scale reference menu.
changes the start and stop values of the stimulus span to the
values of the active marker and the delta reference marker . If
there is no reference marker , the message \NO MARKER DEL T A
SP AN NOT SET" is displayed.
changes the stimulus start value to the stimulus value of the
active marker .
MARKER ! STIMULUS sets the starting stimulus value of a limit line segment using the
active marker . Move the marker to the desired starting stimulus
value before pressing this key , and the marker stimulus value is
entered as the segment start value .
9-22 K ey De nitions

MARKER ! STOP
changes the stimulus stop value to the stimulus value of the
active marker .
MARKER 1 turns on marker 1 and makes it the active marker . The active
marker appears on the display as r . The active marker
stimulus value is displayed in the active entry area, together
with the marker number . If there is a marker turned on, and no
other function is active , the stimulus value of the active marker
can be controlled with the knob , the step keys , or the numeric
keypad. The marker response and stimulus values are displayed
in the upper right-hand corner of the screen.
MARKER 2
turns on marker 2 and makes it the active marker . If another
marker is present, that marker becomes inactive and is
represented on the display as .
MARKER 3
turns on marker 3 and makes it the active marker .
turns on marker 4 and makes it the active marker .
NNNNNNNNNNNNNNNNNNNNNNNNNN
turns on marker 5 and makes it the active marker .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
turns o all the markers and the delta reference marker , as well
as the tracking and bandwidth functions that are accessed with
MKR FCTN key .
key activates a marker if one is not already active , and provides
access to additional marker functions . These can be used
to quickly change the measurement parameters , to search
the trace for speci ed information, and to analyze the trace
statistically .
provides access to the marker mode menu, where several
marker modes can be selected including special markers for
polar and Smith chart formats .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
located under the
key , interpolates between measured
MARKER
points to allow the markers to be placed at any point on the
trace . Displayed marker values are also interpolated. This is the
default marker mode .
channels . Even if the stimulus is uncoupled and two sets of
stimulus values are shown, the markers track the same stimulus
values on each channel as long as they are within the displayed
stimulus range .
places markers only on measured trace points determined by
the stimulus settings .
MARKERS: UNCOUPLED
allows the marker stimulus values to be controlled
independently on each channel.
MAX
moves the active marker to the maximum point on the trace .
is used to de ne the highest frequency at which a calibration
kit standard can be used during measurement calibration. In
waveguide , this is normally the upper cuto frequency of the
standard.
K ey De nitions 9-23
MARKER all OFF · the ·

key provides access to a series of softkey menus for selecting
the parameters or inputs to be measured.
aborts the sweep in progress , then restarts the measurement.
This can be used to update a measurement following an
adjustment of the device under test. When a full two-port
calibration is in use , the
MEASURE RESTART key will initiate
another update of both forward and reverse S-parameter data.
This softkey will also override the test set hold mode , which
inhibits continuous switching of either the test port transfer
switch or step attenuator . The measurement con gurations
which cause this are described in T est Set Attenuator , T est
P ort Transfer Switch, and Doubler Switch Protection, at the
beginning of this section. This softkey will override the test set
hold mode for one measurement.
If the analyzer is taking a number of groups (see Trigger
Menu), the sweep counter is reset at 1. If averaging is on,
MEASURE RESTART resets the sweep-to-sweep averaging and is
e ectively the same as
AVERAGING RESTART . If the sweep
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MEASURE RESTART executes a single
sweep .
displays the trace memory for the active channel. This is the
only memory display mode where the smoothing and gating of
the memory trace can be changed. If no data has been stored in
memory for this channel, a warning message is displayed.
provides access to a series of menus which are used to de ne
and control all stimulus functions other than start, stop , center ,
and span. When the
key is pressed, the stimulus menu is
displayed.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MIDDLE VALUE sets the midpoint for
DELTA LIMITS . It uses the entry controls
to set a speci ed amplitude value vertically centered between
the limits .
MIN moves the active marker to the minimum point on the trace .
MINIMUM FREQUENCY
is used to de ne the lowest frequency at which a calibration
kit standard can be used during measurement calibration. In
waveguide , this must be the lower cuto frequency of the
standard, so that the analyzer can calculate dispersive e ects
OFFSET DELAY ).
leads to the marker search menu, which is used to search the
trace for a particular value or bandwidth.
puts a xed reference marker at the present active marker
position, and makes the xed marker stimulus and response
values at that position equal to zero . All subsequent stimulus
and response values of the active marker are then read out
relative to the xed marker . The xed marker is shown on the
display as a small triangle (delta), smaller than the inactive
marker triangles . The softkey label changes from
MKR ZERO REF = and the notation \REF = " is displayed
at the top right corner of the graticule . Marker zero is canceled
9-24 K ey De nitions

by turning delta mode o in the delta marker menu or turning
all the markers o with the
ALL OFF softkey .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MODIFY [ ]
leads to the modify cal kit menu, where a default cal kit can be
user-modi ed.
MODIFY COLORS present a menu for color modi cation of display elements .
Refer to A djusting Color for information on modifying display
elements .
NETWORK ANALYZER sets the analyzer to network analyzer mode .
NEW SEQ/MODIFY SEQ
activates the sequence edit mode and presents the new/modify
sequence menu with a list of sequences that can be created or
modi ed.
NEWLINE puts a new line command into the display title .
NEXT PAGE
steps forward through a tabular list of data page-by-page .
triggers a user-speci ed number of sweeps , and returns to the
hold mode . This function can be used to override the test set
hold mode , which protects the electro-mechanical transfer
switch and attenuator against continuous switching. This is
explained fully in the T est Set Attenuator description in the
\Application and Operation Concepts" chapter , in this manual.
If averaging is on, the number of groups should be at least
equal to the averaging factor selected to allow measurement of
a fully averaged trace . Entering a number of groups resets the
averaging counter to 1.
NUMBER OF POINTS
is used to select the number of data points per sweep to be
measured and displayed. Using fewer points allows a faster
sweep time but the displayed trace shows less horizontal detail.
Using more points gives greater data density and improved
trace resolution, but slows the sweep and requires more
memory for error correction or saving instrument states .
The possible values that can be entered for number of points
are 3, 11, 26, 51, 101, 201, 401,801, and 1601. The number
of points can be di erent for the two channels if the stimulus
values are uncoupled.
In list frequency sweep , the number of points displayed is the
total number of frequency points for the de ned list (see Sweep
Type Menu).
NUMBER OF READINGS
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
determines the number of measurement/correction iterations
performed on each point in a power meter calibration. This
feature helps eliminate residual power errors after the initial
correction. The amount of residual error is directly proportional
to the magnitude of the initial correction. The user should
initially set the source power so that it is approximately
correct at the device under test. If power uncertainty at the
device under test is expected to be greater than a few dB , it is
recommended that the number of readings be greater than 1.
N
OFFSET
selects the calibration standard load as being o set.
K ey De nitions 9-25

OFFSET DELAY
is used to specify the one-way electrical delay from the
measurement (reference) plane to the standard, in seconds (s).
(In a transmission standard, o set delay is the delay from plane
to plane .) Delay can be calculated from the precise physical
length of the o set, the permittivity constant of the medium,
and the speed of light.
OFFSET LOADS DONE
completes the selection in the O set Load Menu.
is used to specify energy loss , due to skin e ect, along a
one-way length of coax o set. The value of loss is entered as
ohms/nanosecond (or Giga ohms/second) at 1 GHz. (Such losses
are negligible in waveguide , so enter 0 as the loss o set.)
is used to specify the characteristic impedance of the coax
(F or waveguide , the o set impedance should always be assigned
a value equal to the system Z0.)
OMIT ISOLATION
is used to omit the isolation portion of the calibration.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
leads to the series of menus used to perform a high-accuracy
two-port calibration without an S-parameter test set.
This calibration procedure e ectively removes directivity ,
source match, load match, isolation, re ection tracking, and
transmission tracking errors in one direction only . Isolation
correction can be omitted for measurements of devices with
limited dynamic range . (The device under test must be
manually reversed between sweeps to accomplish measurement
of both input and output responses .) The required standards are
a short, an open, a thru, and an impedance-matched load.
This mode does not measure each sweep , but corrects each
point with the data currently in the power meter correction
table .
OP PARMS (MKRS etc) provides a tabular listing on the analyzer display of the key
parameters for both channels . The screen menu is presented to
allow hard copy listings and access new pages of the table . F our
pages of information are supplied. These pages list operating
parameters , marker parameters , and system parameters that
relate to control of peripheral devices rather than selection of
measurement parameters .
de nes the standard type as an open, used for calibrating
re ection measurements . Opens are assigned a terminal
impedance of in nite ohms , but delay and loss o sets may still
be added. Pressing this key also brings up a menu for de ning
the open, including its capacitance .
gets data from an HP-IB device set to the address at which the
analyzer expects to nd a power meter . The data is stored in a
title string. The analyzer must be in system controller or pass
control mode .
PARALL IN BIT NUMBER while creating a sequence , this softkey will insert a command
that selects the single bit (0 to 4) that a sequence will be
looking for from the GPIO bus .
9-26 K ey De nitions

PARALL IN IF BIT H
while creating a sequence , this softkey inserts a command to
jump to another sequence if the single input selected is in a
high state .
PARALL IN IF BIT L
while creating a sequence , this softkey inserts a command to
jump to another sequence if the single input selected is in a low
state .
PARALLEL sets the printer or plotter port to parallel.
toggles the parallel output port between the copy and GPIO
output modes .
PARALLEL OUT ALL
allows you to input a number (0 to 255) in base 10, and outputs
it to the bus as binary , when the parallel port is in GPIO mode .
PAUSE
pauses the sequence so the operator can perform a needed task,
such as changing the DUT , changing the calibration standard, or
other similar task. Press
CONTINUE SEQUENCE when ready .
PAUSE TO SELECT when editing a sequence ,
PAUSE TO SELECT appears when you
DO SEQUENCE . When placed in a sequence , it presents
the menu of up to 6 available sequences (softkeys containing
non-empty sequences). The message \CHOOSE ONE OF THESE
SEQUENCES" is displayed and the present sequence is stopped.
If the operator selects one of the sequences , that sequence is
executed. Any other key can be used to exit this mode . This
function is not executed if used during modify mode and
does nothing when operated manually . This softkey is not
visible on the display , and the function is not available , unless
programmed into analyzer memory .
PEN NUM DATA
selects the number of the pen to plot the data trace . The
default pen for channel 1 is pen number 2, and for channel 2 is
pen number 3.
PEN NUM GRATICULE
selects the number of the pen to plot the graticule . The default
pen for channel 1 is pen number 1, and for channel 2 is pen
number 1.
PEN NUM MARKER selects the number of the pen to plot both the markers and the
marker values . The default pen for channel 1 is pen number 7,
and for channel 2 is pen number 7.
PEN NUM MEMORY
selects the number of the pen to plot the memory trace . The
default pen for channel 1 is pen number 5, and for channel 2 is
pen number 6.
PEN NUM TEXT
selects the number of the pen to plot the text. The default pen
for channel 1 is pen number 7, and for channel 2 is pen
number 7.
PHASE OFFSET adds or subtracts a phase o set that is constant with frequency
(rather than linear). This is independent of
MARKER ! DELAY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ELECTRICAL DELAY .
NNNNNNNNNNNNNNNN
(Option 010 only) displays a Cartesian format of the phase
portion of the data, measured in degrees . This format displays
the phase shift versus frequency .
K ey De nitions 9-27
press · and

PLOT
makes a hard copy plot of one page of the tabular listing on
the display , using a compatible HP plotter connected to the
analyzer through HP-IB . This method is appropriate when speed
of output is not a critical factor .
PLOT DATA ON off
speci es whether the data trace is to be drawn (on) or not
drawn (o ) on the plot.
PLOT GRAT ON off
speci es whether the graticule and the reference line are
to be drawn (on) or not drawn (o ) on the plot. Turning
PLOT GRAT ON and all other elements o is a convenient
way to make preplotted grid forms . However , when data is to
be plotted on a preplotted form,
PLOT GRAT OFF should be
selected.
PLOT MEM ON off speci es whether the memory trace is to be drawn (on) or not
drawn (o ) on the plot. Memory can only be plotted if it is
displayed (refer to \Display Menu" in Chapter 6).
PLOT MKR ON off
speci es whether the markers and marker values are to be
drawn (on) or not drawn (o ) on the plot.
PLOT NAME PLOTFILE
supplies a name for the plot le generated by a PLOT to disk.
Brings up the TITLE FILE MENU .
PLOT SPEED [ ]
toggles between fast and slow speeds .
selects plotting of all displayed text except the marker values ,
PLOT TEXT ON off
softkey labels , and display listings such as the frequency list
table or limit table . (Softkey labels can be plotted under the
control of an external controller . Refer to the Introductory
Programming Guide .)
PLOTTER BAUD RATE
sets the serial port data transmission speed for plots .
sends a page eject command to the plotter .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
sets the HP-IB address the analyzer will use to communicate
with the plotter .
PLTR PORT: DISK
directs plots to the selected disk (internal or external).
directs plots to the HP-IB port and sets the HP-IB address the
analyzer will use to communicate with the plotter .
con gures the analyzer for a plotter that has a parallel
(centronics) interface .
con gures the analyzer for a plotter that has a serial (RS-232)
interface .
PLTR TYPE [PLOTTER] selects a pen plotter such as the HP 7440A, HP 7470A,
HP 7475A, or HP 7550B as the plotter type .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
selects a PCL5 compatible printer , which supports HP-GL/2,
such as the LaserJet III or LaserJet 4 for a monochrome plotter
type , or the DeskJet 1200C for a color plotter type .
displays a polar format. Each point on the polar format
corresponds to a particular value of both magnitude and phase .
determined by its displacement from the center (which has zero
9-28 K ey De nitions

value), and the phase by the angle counterclockwise from the
positive x-axis . Magnitude is scaled in a linear fashion, with the
value of the outer circle usually set to a ratio value of 1. Since
there is no frequency axis , frequency information is read from
the markers .
POLAR MKR MENU leads to a menu of special markers for use with a polar format.
PORT EXTENSIONS
goes to the reference plane menu, which is used to extend the
apparent location of the measurement reference plane or input.
PORT PWR [COUPLED]
is used to set the same power levels at each port.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
allows you to set di erent power levels at each port.
NNNNNNNNNNNNNNNNN
POWER
makes power level the active function and sets the RF
output power level of the analyzer's internal source . The
analyzer will detect an input power overload at any of the
three receiver inputs , and automatically reduce the output
power of the source to 85 dBm. This is indicated with the
message \O VERLO AD ON INPUT (R, A, B)." In addition,
the annotation \P# " appears at the left side of the display .
When this occurs , set the power to a lower level, and toggle
SOURCE PWR on OFF . If power meter cal is on, cal power is
the active entry .
POWER: FIXED
sets the external LO xed power .
sets the external LO power sweep .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
brings up the segment modify menu and segment edit (power
loss) menu explained in the following pages . This softkey is
intended for use when the power output is being sampled
by a directional coupler or power splitter . In the case of the
directional coupler , enter the power loss caused by the coupled
arm. Refer to P ower Loss F eature on a previous page .
This feature may be used to compensate for attenuation
non-linearities in either a directional coupler or a power splitter .
Up to 12 segments may be entered, each with a di erent
frequency and power loss value .
NNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNN
POWER MTR toggles between
438A/437 . These power meters are
HP-IB compatible with the analyzer . The model number in the
softkey label must match the power meter to be used.
turns on a power sweep mode that is used to characterize
power-sensitive circuits . In this mode , power is swept at a
single frequency , from a start power value to a stop power
value , selected using the
STOP
keys and the entry
block. This feature is convenient for such measurements as
gain compression or A GC (automatic gain control) slope . T o
set the frequency of the power sweep , use
CW FREQ in the
stimulus menu. Refer to the User's Guide for an example of a
gain compression measurement.
mode .
K ey De nitions 9-29
436A or · and
![Page 516 of the Agilent Technologies 8753D user manual — PRNTR TYPE [THINKJET]](/cdn/reader/36/366f615c4db1b577727decdae885e4a4811cfac039b6c91f50ac1e89d34afa65/images/page-516.webp?w=400)
In power sweep , the entered sweep time may be automatically
changed if it is less than the minimum required for the current
con guration (number of points , IF bandwidth, averaging, etc .).
presents a menu to select a factory or user de ned preset state .
PRESET: FACTORY
is used to select the preset conditions de ned by the factory .
is used to select a preset condition de ned by the user . This
is done by saving a state in a register under
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
and naming the register UPRESET . When
key will bring up the state of the
UPRESET register .
PREVIOUS PAGE
steps backward through a tabular list of data page-by-page .
when displaying list values , prints the entire list in color . When
displaying operating parameters , prints all but the last page in
color . The data is sent to the printer as ASCII text rather than
as raster graphics , which causes the printout to be faster .
PRINT ALL MONOCHROME
when displaying list values , prints the entire list in
monochrome . When displaying operating parameters , prints
all but the last page in monochrome . The data is sent to the
printer as ASCII text rather than as raster graphics , which
causes the printout to be faster .
PRINT: COLOR
sets the print command to default to a color printer . The
printer output is always in the analyzer default color values .
This command does not work with a black and white printer .
PRINT COLOR
prints the displayed measurement results in color .
is used to select the print colors menu.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PRINT: MONOCHROME
sets the print command to default to a black and white printer .
prints the displayed measurement results in black and white .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
prints any sequence currently in memory to a compatible
printer .
PRINTER BAUD RATE
sets the serial port data transmission speed for prints .
sends a conditional form feed to the printer .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PRINTER PORT
sets the HP-IB address the analyzer will use to communicate
with the printer .
PRNTR TYPE [DESKJET]
sets the printer type to the DeskJet series .
sets the printer type to Epson compatible printers , which
support the Epson ESC/P2 printer control language .
sets the printer type to the LaserJet series .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PRNTR TYPE [PAINTJET]
sets the printer type to the P aintJet.
sets the printer type to the ThinkJet or QuietJet.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PWR LOSS on OFF
turns on or o power loss correction. P ower loss correction
should be used when the power output is measured by a
9-30 K ey De nitions
· · underlined, the

directional coupler . Enter the power loss caused by the coupled
LOSS/SENSR LISTS softkey submenus described
below .
PWR RANGE AUTO man toggles the power range mode between auto and manual. A uto
mode selects the power range based on the power selected.
Manual mode limits power entry to within the selected range .
PWRMTR CAL [ ]
leads to the power meter calibration menu which provides two
types of power meter calibration, continuous and single-sample .
turns o power meter calibration.
NNNNN
R
measures the absolute power amplitude at input R.
converts the active marker values into rectangular form. The
complex impedance values of the active marker are displayed in
terms of resistance , reactance , and equivalent capacitance or
inductance . This is the default Smith chart marker .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RANGE 0 -15 TO +10 selects power range 0 when in manual power range .
selects power range 1 when in manual power range .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RANGE 2 -35 TO -10
selects power range 2 when in manual power range .
selects power range 3 when in manual power range .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RANGE 4 -55 TO -30
selects power range 4 when in manual power range .
selects power range 5 when in manual power range .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
selects power range 6 when in manual power range .
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RANGE 7 -85 TO -60
selects power range 7 when in manual power range .
speci es whether or not to store the raw data (ratioed and
RAW ARRAY on OFF
averaged) on disk with the instrument state .
selects whether sampler and attenuator o sets are ON or OFF .
RAW OFFSET On Off
By selecting raw o sets OFF , a full two port error correction
can be performed without including the e ects of the o sets .
It also saves substantial time at recalls and during frequency
changes . Raw o sets follow the channel coupling. This softkey
is used with \T ake4" mode . See \Example 2E" in Chapter 2 of
the HP 8753D Programmer's Guide.
Re/Im MKR when in the smith marker menu,
Re/Im MKR displays the
values of the active marker on a Smith chart as a real and
imaginary pair . The complex data is separated into its real part
and imaginary part. The rst marker value given is the real
part M cos , and the second value is the imaginary part M sin
, where M = magnitude .
When in the polar marker menu,
Re/Im MKR displays the
values of the active marker as a real and imaginary pair . The
complex data is separated into its real part and imaginary
part. The rst marker value given is the real part M cos ,
and the second value is the imaginary part M sin , where
M = magnitude .
K ey De nitions 9-31
arm with the · R+jX MKR

READ FILE TITLES
searches the directory of the disk for le names recognized
as belonging to an instrument state , and displays them in the
softkey labels . No more than ve titles are displayed at one
time . If there are more than ve , repeatedly pressing this key
causes the next ve to be displayed. If there are fewer than
ve , the remaining softkey labels are blanked.
READ SEQ FILE TITLS
is a disk le directory command. Pressing this softkey will read
the rst six sequence titles and display them in the softkey
labels as described in Loading a Sequence When the Title Is
Not Known. These sequences can then be loaded into internal
memory .
READ SEQ FILE TITLS is pressed again, the next six
sequence titles on the disk will be displayed. T o read the
contents of the disk starting again with the rst sequence:
remove the disk, reinsert it into the drive , and press
READ SEQ FILE TITLS .
displays only the real (resistive) portion of the measured data
on a Cartesian format. This is similar to the linear magnitude
format, but can show both positive and negative values . It is
primarily used for analyzing responses in the time domain, and
also to display an auxiliary input voltage signal for service
purposes .
RECALL CAL PORT 1
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Press this key after selecting the le associated with port 1
error correction for adapter removal calibration.
RECALL CAL PORT 2
Press this key after selecting the le associated with port 2
error correction for adapter removal calibration.
RECALL COLORS
recalls the previously saved modi ed version of the color set.
This key appears only when a color set has been saved.
RECALL KEYS MENU
provides access to the recall keys menu where speci c registers
can be recalled.
RECALL KEYS on OFF presents the recall keys menu as the initial menu when
has been pressed.
RECALL REG1
recalls the instrument state saved in register 1.
recalls the instrument state saved in register 2.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
recalls the instrument state saved in register 3.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RECALL REG4
recalls the instrument state saved in register 4.
recalls the instrument state saved in register 5.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RECALL REG6
recalls the instrument state saved in register 6.
recalls the instrument state saved in register 7.
N
RECALL STATE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
is used in conjunction with sequencing, to return the
instrument to the known preset state without turning o
the sequencing function. This is not the same as pressing
sequencing activities are not changed.
9-32 K ey De nitions
If · · the

RECEIVER CAL
provides access to the Receiver Cal Menu.
sets the position of the reference line on the graticule of a
Cartesian display , with 0 the bottom line of the graticule and 10
the top line . It has no e ect on a polar or Smith display . The
reference position is indicated with a small triangle just outside
the graticule , on the left side for channel 1 and the right side
for channel 2.
REFERENCE VALUE changes the value of the reference line , moving the
measurement trace correspondingly . In polar and Smith chart
formats , the reference value is the same as the scale , and is the
value of the outer circle .
, the complex re ection
coecient (magnitude and phase) of the test device input.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
de nes the measurement as S
, the complex re ection
coecient (magnitude and phase) of the output of the device
under test.
REFLECT AND LINE measures the re ection and thru paths of the current
calibration standard.
REFLECTION leads to the re ection calibration menu.
completes the adapter removal procedure , removing the e ects
of the adapter being used.
RENAME FILE
allows you to change the name of a le that has already been
saved.
RESET COLOR resets the color being modi ed to the default color .
NNNNNNNNNNNNNNNNNNNNNNNNNN
RESPONSE When in the specify class more menu,
RESPONSE is used to
enter the standard numbers for a response calibration. This
calibration corrects for frequency response in either re ection
or transmission measurements , depending on the parameter
being measured when a calibration is performed. (F or default
kits , the standard is either the open or short for re ection
measurements , or the thru for transmission measurements .)
When in the response cal menu,
RESPONSE leads to the
frequency response calibration. This is the simplest and
fastest accuracy enhancement procedure , but should be
used when extreme accuracy is not required. It e ectively
removes the frequency response errors of the test setup for
re ection or transmission measurements .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RESPONSE & ISOL'N When in the specify class more menu,
is used to enter the standard numbers for a response and
isolation calibration. This calibration corrects for frequency
response and directivity in re ection measurements ,
or frequency response and isolation in transmission
measurements .
When in the response and isolation menu,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RESPONSE & ISOL'N leads to the menus used to perform
a response and isolation measurement calibration, for
measurement of devices with wide dynamic range . This
K ey De nitions 9-33

procedure e ectively removes the same frequency response
errors as the response calibration. In addition, it e ectively
removes the isolation (crosstalk) error in transmission
measurements or the directivity error in re ection
measurements . As well as the devices required for a simple
response calibration, an isolation standard is required.
The standard normally used to correct for isolation is an
impedance-matched load (usually 50 or 75 ohms). Response
and directivity calibration procedures for re ection and
transmission measurements are provided in the following
pages .
RESTORE DISPLAY
turns o the tabular listing and returns the measurement
display to the screen.
RESUME CAL SEQUENCE
eliminates the need to restart a calibration sequence that was
interrupted to access some other menu. This softkey goes back
to the point where the calibration sequence was interrupted.
REV ISOL'N ISOL'N STD
measures the reverse isolation of the calibration standard.
REV MATCH (Label Class)
lets you enter a label for the reverse match class . The label
appears during a calibration that uses this class .
REV MATCH (Specify Class)
speci es which standards are in the reverse match class in the
calibration kit.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
REV MATCH THRU is used to enter the standard numbers for the reverse match
(thru) calibration. (F or default kits , this is the thru.)
REV TRANS (Label Class)
lets you enter a label for the reverse transmission class . The
label appears during a calibration that uses this class .
REV TRANS (Specify Class)
speci es which standards are in the reverse transmission class
in the calibration kit.
REV TRANS THRU is used to enter the standard numbers for the reverse
transmission (thru) calibration. (F or default kits , this is the
thru.)
NNNNNNNNNNNNNNNNNNNNNN
RF > LO
adjusts the source frequency higher than the LO by the amount
of the LO (within the limits of the analyzer).
RF < LO
adjusts the source frequency lower than the LO by the amount
of the LO (within the limits of the analyzer).
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RIGHT LOWER
draws a quarter-page plot in the lower right quadrant of the
page .
RIGHT UPPER
draws a quarter-page plot in the upper right quadrant of the
page .
ROUND SECONDS
resets the seconds counter to zero in real-time clock.
S PARAMETERS
presents the S-parameter menu, which is used to de ne
the input ports and test set direction for S-parameter
measurements .
S11 1-PORT provides a measurement calibration for re ection-only
measurements of one-port devices or properly terminated
two-port devices , at port 1 of an S-parameter test set or the test
port of a transmission/re ection test set.
9-34 K ey De nitions

S11A
is used to enter the standard numbers for the rst class
required for an S
1-port calibration. (F or default cal kits , this
is the open.)
S11B
is used to enter the standard numbers for the second class
required for an S
1-port calibration. (F or default cal kits , this
is the short.)
S11C
is used to enter the standard numbers for the third class
required for an S
1-port calibration. (F or default kits , this is
the load.)
S11 REFL SHORT
measures the short circuit TRL/LRM calibration data for
PORT 1.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
S22 1-PORT provides a measurement calibration for re ection-only
measurements of one-port devices or properly terminated
two-port devices , at port 2 of an S-parameter test set or the test
port of a transmission/re ection test set.
S22A
is used to enter the standard numbers for the rst class
required for an S
1-port calibration. (F or default cal kits , this
is the open.)
S22B
is used to enter the standard numbers for the second class
required for an S
1-port calibration. (F or default cal kits , this
is the short.)
S22C
is used to enter the standard numbers for the third class
required for an S
1-port calibration. (F or default kits , this is
the load.)
S22 REFL SHORT
measures the short circuit TRL/LRM calibration data for
PORT 2.
SAMPLR COR on OFF selects whether sampler correction is on or o .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
saves the modi ed version of the color set.
provides access to all the menus used for saving and recalling
instrument states in internal memory and for storing to , or
loading from, external disk. This includes the menus used to
de ne titles for internal registers and external disk les , to
de ne the content of disk les , to initialize disks for storage ,
and to clear data from the registers or purge les from disk.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
stores the user-modi ed or user-de ned kit into memory , after
it has been modi ed.
SAVE USING ASCII selects ASCII format for data storage to disk.
selects binary format for data storage .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SCALE/DIV
changes the response value scale per division of the displayed
trace . In polar and Smith chart formats , this refers to the full
scale value at the outer circumference , and is identical to
reference value .
SCALE PLOT [ ] toggles between two selections for plot scale , FULL and GRA T .
SCALE PLOT [FULL]
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
is the normal scale selection for plotting on blank paper . It
includes space for all display annotations such as marker
values , stimulus values , etc . The entire display ts within the
K ey De nitions 9-35

user-de ned boundaries of P1 and P2 on the plotter , while
maintaining the exact same aspect ratio as the display .
SCALE PLOT [GRAT]
expands or reduces the horizontal and vertical scale so that the
lower left and upper right graticule corners exactly correspond
to the user-de ned P1 and P2 scaling points on the plotter . This
is convenient for plotting on preprinted rectangular or polar
forms (for example , on a Smith Chart).
] makes scale per division the active function. A menu is
displayed that is used to modify the vertical axis scale and
the reference line value and position. In addition this menu
provides electrical delay o set capabilities for adding or
subtracting linear phase to maintain phase linearity .
SEARCH LEFT
searches the trace for the next occurrence of the target value
to the left.
SEARCH RIGHT searches the trace for the next occurrence of the target value
to the right.
SEARCH: MAX
moves the active marker to the maximum point on the trace .
moves the active marker to the minimum point on the trace .
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEARCH: OFF
turns o the marker search function.
SEGMENT
speci es which limit segment in the table is to be modi ed.
A maximum of three sets of segment values are displayed at
one time , and the list can be scrolled up or down to show
other segment entries . Use the entry block controls to move
the pointer > to the required segment number . The indicated
segment can then be edited or deleted. If the table of limits is
designated \EMPTY ," new segments can be added using the
NNNNNNNNNNNNNN
EDIT softkey .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEGMENT: CENTER
sets the center frequency of a subsweep in a list frequency
sweep .
SEGMENT: SPAN
sets the frequency or power span of a subsweep about a
speci ed center frequency .
SEGMENT: START
sets the start frequency of a subsweep .
SEGMENT: STOP
sets the stop frequency of a subsweep .
leads to the select quadrant menu, which provides the
capability of drawing quarter-page plots . This is not used for
printing.
provides access to the select disk menu.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SELECT LETTER
The active entry area displays the letters of the alphabet, digits
0 through 9, and mathematical symbols . T o de ne a title , rotate
the knob until the arrow " points at the rst letter , then press
SELECT LETTER . Repeat this until the complete title is de ned,
for a maximum of 50 characters . As each character is selected,
it is appended to the title at the top of the graticule .
SELF DIAGNOSE
prompts the analyzer to run a series of tests to determine a
problem.
9-36 K ey De nitions
· ADD or

accesses a series of sequencing menus . These allow you to
create , modify , and store up to 6 sequences which can be run
automatically .
SEQUENCE 1 SEQ1
activates editing mode for the segment titled \SEQ1" (default
title).
SEQUENCE 2 SEQ2
activates editing mode for the segment titled \SEQ2" (default
title).
SEQUENCE 3 SEQ3
activates editing mode for the segment titled \SEQ3" (default
title).
SEQUENCE 4 SEQ4
activates editing mode for the segment titled \SEQ4" (default
title).
SEQUENCE 5 SEQ5
activates editing mode for the segment titled \SEQ5" (default
title).
SEQUENCE 6 SEQ6
activates editing mode for the segment titled \SEQ6" (default
title).
SEQUENCE FILENAMING
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
accesses a lenaming menu which is used to automatically
increment or decrement the name of a le that is generated by
the network analyzer during a SEQUENCE.
SERVICE MENU
leads to a series of service and test menus described in detail in
the On-Site System Service Manual.
SERVICE MODES
a collection of common modes used for troubleshooting.
goes to the address menu, which is used to set the HP-IB
address of the analyzer , and to display and modify the
addresses of peripheral devices in the system.
allows you to set the analyzer's internal clock.
NNNNNNNNNNNNNNNNNNNNNNN
SET DAY
allows you to set the day in the analyzer's internal clock.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
(Option 010 only) changes the frequency sweep to harmonic
intervals to accommodate time domain low-pass operation
(option 010). If this mode is used, the frequencies must be set
before calibration.
SET HOUR allows you to set the hour in the analyzer's internal clock.
allows you to set the minutes in the analyzer's internal clock.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SET MONTH
allows you to set the month in the analyzer's internal clock.
sets the measurement reference plane to the TRL/LRM
REFLECT standard.
standard.
SET YEAR allows you to set the year in the analyzer's internal clock.
sets the characteristic impedance used by the analyzer in
calculating measured impedance with Smith chart markers and
conversion parameters . Characteristic impedance must be set
correctly before calibration procedures are performed.
K ey De nitions 9-37

SHORT
de nes the standard type as a short, for calibrating re ection
measurements . Shorts are assigned a terminal impedance of
0 ohms , but delay and loss o sets may still be added.
SHOW MENUS
used to display a speci c menu prior to a pause statement.
takes one sweep of data and returns to the hold mode .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
sets the limits at a single stimulus point. If limit lines are on,
the upper limit value of a single point limit is displayed as n 9,
and the lower limit is displayed as n 8. A limit test at a single
point not terminating a at or sloped line tests the nearest
actual measured data point. A single point limit can be used as
a termination for a at line or sloping line limit segment. When
a single point terminates a sloping line or when it terminates a
at line and has the same limit values as the at line , the single
point is not displayed as n 9 and n 8. The indication for a sloping
line segment in the displayed table of limits is SP .
SINGLE SEG SWEEP
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
enables a measurement of a single segment of the frequency
list, without loss of calibration. The segment to be measured is
selected using the entry block.
In single segment mode , selecting a measurement calibration
will force the full list sweep before prompting for calibration
standards . The calibration will then be valid for any single
segment.
If an instrument state is saved in memory with a single-segment
trace , a recall will re-display that segment while also recalling
the entire list.
NNNNNNNNNNNNNNNNNNNNNN
SLIDING de nes the load as a sliding load. When such a load is measured
during calibration, the analyzer will prompt for several load
positions , and calculate the ideal load value from it.
SLOPE
compensates for power loss versus the frequency sweep , by
sloping the output power upwards proportionally to frequency .
Use this softkey to enter the power slope in dB per GHz of
sweep .
toggles the power slope function on or o . With slope on, the
output power increases with frequency , starting at the selected
power level.
SLOPING LINE
de nes a sloping limit line segment that is linear with frequency
or other stimulus value , and is continuous to the next stimulus
value and limit. If a sloping line is the nal segment it becomes
a at line terminated at the stop stimulus . A sloping line
segment is indicated as SL on the displayed table of limits .
SMITH CHART
displays a Smith chart format. This is used in re ection
measurements to provide a readout of the data in terms of
impedance .
SMITH MKR MENU
leads to a menu of special markers for use with a Smith chart
format.
SMOOTHING APERTURE lets you change the value of the smoothing aperture as a
percent of the span. When smoothing aperture is the active
9-38 K ey De nitions
SLOPE on OFF

function, its value in stimulus units is displayed below its
percent value in the active entry area.
Smoothing aperture is also used to set the aperture for group
delay measurements . Note that the displayed smoothing
aperture is not the group delay aperture unless smoothing is on.
turns the smoothing function on or o for the active channel.
When smoothing is on, the annotation \Smo" is displayed in the
status notations area.
SOURCE PWR on OFF turns the source power on or o . Use this key to restore
power after a power trip has occurred. (See the
description.)
SPACE
inserts a space in the title .
is used, along with the
key , to de ne the frequency
CENTER
range of the stimulus . When the
key is pressed it
becomes the active function. The value is displayed in the
active entry area, and can be changed with the knob , step keys ,
or numeric keypad.
sets the frequency or power span of a subsweep about a
speci ed center frequency .
presents the special function menu.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SPECIFY CLASS
leads to the specify class menu. After the standards are
modi ed, use this key to specify a class to consist of certain
standards .
SPECIFY CLASS DONE nishes the specify class function and returns to the modify cal
kit menu.
SPECIFY GATE (Option 010 only) is used to specify the parameters of the gate .
SPECIFY OFFSET
allows additional speci cations for a user-de ned standard.
F eatures speci ed in this menu are common to all ve types of
standards .
SPLIT DISP on OFF toggles between a full-screen single graticule display of one
or both channels , and a split display with two half-screen
graticules one above the other . The split display can be
used in conjunction with
DUAL CH ON in the display menu
to show the measured data of each channel simultaneously
on separate graticules . In addition, the stimulus functions
of the two channels can be controlled independently using
COUPLED CH ON in the stimulus menu. The markers can
also be controlled independently for each channel using
MARKERS: UNCOUPLED in the marker mode menu.
selects whether spur avoidance is ON or OFF . Selecting spur
SPUR AVOID On Off
avoidance OFF , along with selecting raw o sets OFF , saves
substantial time at recalls and during frequency changes . Spur
avoidance is always coupled between channels .
is used to de ne the start frequency of a frequency range .
key is pressed it becomes the active function.
K ey De nitions 9-39

The value is displayed in the active entry area, and can be
changed with the knob , step keys , or numeric keypad.
calculates and displays the mean, standard deviation, and
peak-to-peak values of the section of the displayed trace
between the active marker and the delta reference marker .
If there is no delta reference , the statistics are calculated for
the entire trace . A convenient use of this feature is to nd
the peak-to-peak value of passband ripple without searching
separately for the maximum and minimum values .
to de ne the span. F or polar and Smith chart formats the
statistics are calculated using the rst value of the complex pair
(magnitude , real part, resistance , or conductance).
After each standard is de ned, including o sets , press
STD DONE (DEFINED) to terminate the standard de nition.
returns to the de ne standard menu.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
STD OFFSET DONE
is used to end the specify o set sequence .
is used to specify the type of calibration device being measured.
STD TYPE:
de nes the standard type to be a load, but with an arbitrary
impedance (di erent from system Z0).
de nes the standard type as a transmission line of speci ed
length, for calibrating transmission measurements .
de nes the standard type as a load (termination). Loads
are assigned a terminal impedance equal to the system
characteristic impedance ZO , but delay and loss o sets may still
be added. If the load impedance is not ZO , use the arbitrary
impedance standard de nition.
de nes the standard type as an open used for calibrating
re ection measurements . Opens are assigned a terminal
impedance of in nite ohms , but delay and loss o sets may still
be added. Pressing this key also brings up a menu for de ning
the open, including its capacitance .
de nes the standard type as a short used for calibrating
re ection measurements . Shorts are assigned a terminal
impedance of 0 ohms , but delay and loss o sets may still be
added.
STEP SIZE is used to specify the subsweep in frequency steps instead
of number of points . Changing the start frequency , stop
frequency , span, or number of points may change the step size .
Changing the step size may change the number of points and
stop frequency in start/stop/step mode; or the frequency span
in center/span/step mode . In each case , the frequency span
becomes a multiple of the step size .
sets the starting stimulus value of a segment, using entry block
controls . The ending stimulus value of the segment is de ned
by the start of the next line segment. No more than one
segment can be de ned over the same stimulus range .
9-40 K ey De nitions

STIMULUS OFFSET
adds or subtracts an o set in stimulus value . This allows limits
already de ned to be used for testing in a di erent stimulus
range . Use the entry block controls to specify the o set
required.
is used to de ne the stop frequency of a frequency range .
key is pressed it becomes the active function.
The value is displayed in the active entry area, and can be
changed with the knob , step keys , or numeric keypad.
STOP
sets the stop frequency of a subsweep .
presents the store sequence to disk menu with a list of
sequences that can be stored.
is used to set the frequency of the LO source to sweep .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SWEEP TIME [ ]
toggles between automatic and manual sweep time .
presents the sweep type menu, where one of the available
types of stimulus sweep can be selected.
NNNNNNNNNN
reformats a re ection measurement into its equivalent SWR
(standing wave ratio) value . SWR is equivalent to (1+)/(1),
where is the re ection coecient. Note that the results are
valid only for re ection measurements . If the SWR format is
used for measurements of S
the results are not valid.
presents the system menu.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
is the mode used when peripheral devices are to be used and
there is no external controller . In this mode , the analyzer can
directly control peripherals (plotter , printer , disk drive , or
power meter). System controller mode must be set in order for
the analyzer to access peripherals from the front panel to plot,
print, store on disk, or perform power meter functions , if there
is no other controller on the bus .
The system controller mode can be used without knowledge
of HP-IB programming. However , the HP-IB address must be
entered for each peripheral device .
This mode can only be selected manually from the analyzer's
front panel, and can be used only if no active computer
controller is connected to the system through HP-IB . If you
try to set system controller mode when another controller
is present, the message ANOTHER SYSTEM CONTROLLER ON
HP-IB BUS is displayed. Do not attempt to use this mode for
programming.
TAKE CAL SWEEP
Each data point is measured during the initial sweep and the
correction data is placed in the power meter correction table .
This provides data usable in the
ONE SWEEP mode .
TAKE RCVR CAL SWEEP
executes a receiver calibration.
is the mode normally used for remote programming of the
analyzer . In this mode , the analyzer and all peripheral devices
are controlled from the external controller . The controller can
command the analyzer to talk, and the plotter or other device
to listen. The analyzer and peripheral devices cannot talk
K ey De nitions 9-41

directly to each other unless the computer sets up a data path
between them.
This mode allows the analyzer to be either a talker or a listener ,
as required by the controlling computer for the particular
operation in progress .
A talker is a device capable of sending out data when it is
addressed to talk. There can be only one talker at any given
time . The analyzer is a talker when it sends information over
the bus .
A listener is a device capable of receiving data when it is
addressed to listen. There can be any number of listeners at
any given time . The analyzer is a listener when it is controlled
over the bus by a computer .
NNNNNNNNNNNNNNNNNNN
TARGET
makes target value the active function, and places the active
marker at a speci ed target point on the trace . The default
target value is 3 dB . The target menu is presented, providing
search right and search left options to resolve multiple
solutions .
F or relative measurements , a search reference must be de ned
with a delta marker or a xed marker before the search is
activated.
TERMINAL IMPEDANCE is used to specify the (arbitrary) impedance of the standard, in
ohms .
TEST OPTIONS is used to set con gurations before running the service tests .
TESTPORT 1 2
is used to direct the RF power to port 1 or port 2. (F or non-S
parameter inputs only .)
TESTSET I/O FWD
is used to support specialized test sets , such as a testset that
measures duplexers . It allows you to set three bits (D1, D2,
and D3) to a value of 0 to 7, and outputs it as binary from the
rear panel testset connector . It tracks the coupling ag, so if
coupling is on, and FWD channel 1 is the active channel, FWD
channel 2 will be set to the same value .
TESTSET I/O REV is used to support specialized testsets , such as a testset that
measures duplexers . It allows you to set three bits (D1, D2,
and D3) to a value of 0 to 7, and outputs it as binary from the
rear panel testset connector . It tracks the coupling ag, so if
coupling is on, and REV channel 1 is the active channel, REV
channel 2 will be set to the same value .
TESTSET SW XXXX toggles the internal solid state switch from a hold mode , to a
continuously switching mode , or to a number of sweeps mode
when full 2-port correction is enabled. Use for fast 2-port
calibration.
NNNNNNNNNNNNNNNN
TESTS presents the service test menu.
TEXT
selects all the non-data text for color modi cation. F or
example: operating parameters .
TEXT [ ]
brings up the color de nition menu. The default color for text
is black.
9-42 K ey De nitions

THRU a calibration standard type .
measures all four S-parameters in a TRL/LRM calibration.
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
turns the time stamp function on or o .
TIME STAMP on OFF
TINT
adjusts the continuum of hues on the color wheel of the chosen
attribute . See A djusting Color for an explanation of using this
softkey for color modi cation of display attributes .
TITLE
presents the title menu in the softkey labels area and the
character set in the active entry area. These are used to label
the active channel display . A title more menu allows up to four
values to be included in the printed title; active entry , active
marker amplitude , limit test results , and loop counter value .
TITLE SEQUENCE
allows the operator to rename any sequence with an eight
character title . All titles entered from the front panel must
begin with a letter , and may only contain letters and numbers .
A procedure for changing the title of a sequence is provided at
the beginning of this chapter .
TITLE TO MEMORY
moves the title string data obtained with the
P MTR/HPIB TO TITLE command into a data array .
TITLE TO MEMORY strips o leading characters that are not
numeric , reads the numeric value , and then discards everything
else . The number is converted into analyzer internal format,
and is placed into the real portion of the memory trace at:
Display point = total points - 1 - loop counter
If the value of the loop counter is zero , then the title number
goes in the last point of memory . If the loop counter is greater
than or equal to the current number of measurement points ,
the number is placed in the rst point of memory . A data to
memory command must be executed before using the title to
memory command.
TITLE TO P MTR/HPIB
outputs a title string to any device with an HP-IB address
that matches the address set with the analyzer
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SET ADDRESSES
softkey is generally used for two purposes:
Sending a title to a printer when a CR-LF is not desired.
Sending commands to an HP-IB device .
TITLE TO PERIPHERAL
outputs a title string to any device with an HP-IB address
that matches the address set with the analyzer
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SPECIAL FUNCTIONS
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PERIPHERAL HPIB ADDR commands . This
softkey is generally used for two purposes:
Sending a title to a printer when a CR-LF is not desired.
Sending commands to an HP-IB device .
outputs a title string to any device with an HP-IB address
that matches the address set with the analyzer
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SET ADDRESSES
is generally used for two purposes:
K ey De nitions 9-43
· ·

Sending a title to a printer for data logging or documentation
purposes .
Sending commands to a printer or other HP-IB device .
is used in conjunction with other search features to track the
search with each new sweep . Turning tracking on makes the
analyzer search every new trace for the speci ed target value
and put the active marker on that point. If bandwidth search
is on, tracking searches every new trace for the speci ed
bandwidth, and repositions the dedicated bandwidth markers .
When tracking is o , the target is found on the current sweep
and remains at the same stimulus value regardless of changes in
trace response value with subsequent sweeps .
A maximum and a minimum point can be tracked
simultaneously using two channels and uncoupled markers .
TRANS DONE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
goes back to the two-port cal menu when transmission
measurements are nished.
, the complex forward
transmission coecient (magnitude and phase) of the test
device .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
, the complex reverse
transmission coecient (magnitude and phase) of the test
device .
TRANSFORM MENU (Option 010 only) leads to a series of menus that transform the
measured data from the frequency domain to the time domain.
(Option 010 only) switches between time domain transform on
and o .
TRANSMISSION leads to the transmission menu.
TRIGGER MENU
presents the trigger menu, which is used to select the type and
number of the sweep trigger .
TRIGGER: TRIG OFF
turns o external trigger mode .
leads to the TRL*/LRM* 2-port calibration menu.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TRL/LRM OPTION
selects the TRL/LRM Option Menu.
is used to enter the standard numbers for the TRL LINE or
MA TCH class .
TRL THRU is used to enter the standard numbers for the TRL THRU class .
TRL REFLECT
is used to enter the standard numbers for the TRL REFLECT
class .
TTL OUT HIGH sets the TTL output (TEST SEQ BNC) on the back of the
analyzer high.
TTL OUT LOW
sets the TTL output (TEST SEQ BNC) on the back of the
analyzer low .
TUNED RECEIVER
sets the analyzer to function as a tuned receiver only , disabling
the source .
9-44 K ey De nitions
TRACKING on OFF · TRANSFORM on OFF

UNCOUPLED allows the marker stimulus values to be controlled
independently on each channel.
sends the sum frequency of the RF and LO to the R channel.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
UPPER LIMIT
sets the upper limit value for the start of the segment. If a
lower limit is speci ed, an upper limit must also be de ned. If
no upper limit is required for a particular measurement, force
the upper limit value out of range (for example +500 dB).
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LOWER LIMIT is pressed, all the
segments in the table are displayed in terms of upper and lower
limits , even if they were de ned as delta limits and middle
value .
If you attempt to set an upper limit that is lower than the lower
limit, or vice versa, both limits will be automatically set to the
same value .
USE MEMORY ON off (Option 010 only) remembers a speci ed window pulse width
(or step rise time) di erent from the standard window values .
A window is activated only for viewing a time domain response ,
and does not a ect a displayed frequency domain response .
USE PASS CONTROL
lets you control the analyzer with the computer over HP-IB as
with the talker/listener mode , and also allows the analyzer to
become a controller in order to plot, print, or directly access
an external disk. During this peripheral operation, the host
computer is free to perform other internal tasks that do not
require use of the bus (the bus is tied up by the network
analyzer during this time).
The pass control mode requires that the external controller is
programmed to respond to a request for control and to issue
a take control command. When the peripheral operation is
complete , the analyzer passes control back to the computer .
Refer to the \HP-IB Programming Reference" and \HP-IB
Programming Examples" chapters in the HP 8753D Network
Analyzer Programmer's Guide for more information.
In general, use the talker/listener mode for programming the
analyzer unless direct peripheral access is required.
USE SENSOR A/B
selects the A or B power sensor calibration factor list for use in
power meter calibration measurements .
USER
is used to select the preset condition de ned by the user .
is used to de ne kits other than those o ered by
Hewlett-P ackard.
VELOCITY FACTOR Enters the velocity factor used by the analyzer to calculate
equivalent electrical length in distance-to-fault measurements
using the time domain option. V alues entered should be less
than 1.
V elocity factor is the ratio of the velocity of wave propagation
in a coaxial cable to the velocity of wave propagation in free
space . Most cables have a relative velocity of about 0.66 the
K ey De nitions 9-45
When · UPPER LIMIT or

speed in free space . This velocity depends on the relative
permittivity of the cable dielectric (
V el ocity F actor =
VIEW MEASURE
toggles to become view setup when the analyzer is in frequency
o set mode .
VOLUME NUMBER speci es the number of the disk volume to be accessed. In
general, all 3.5 inch oppy disks are considered one volume
(volume 0). F or hard disk drives , such as the HP 9153A
(Winchester), a switch in the disk drive must be set to de ne
the number of volumes on the disk. F or more information,
refer to the manual for the individual hard disk drive .
WAIT x pauses the execution of subsequent sequence commands for x
number of seconds . T erminate this command with
.
Entering a 0 in wait x causes the instrument to wait for prior
sequence command activities to nish before allowing the
next command to begin. The wait 0 command only a ects
the command immediately following it, and does not a ect
commands later in the sequence .
WARNING
selects the warning annotation for color modi cation.
brings up the color de nition menu. The warning annotation
default color is black.
WAVEGUIDE de nes the standard (and the o set) as rectangular waveguide .
This causes the analyzer to assume a dispersive delay (see
OFFSET DELAY above).
WAVEGUIDE DELAY
applies a non-linear phase shift for use with electrical delay
which follows the standard dispersive phase equation for
rectangular waveguide . When
WAVEGUIDE DELAY is pressed,
the active function becomes the W A VEGUIDE CUTOFF
frequency , which is used in the phase equation. Choosing a
Start frequency less than the Cuto frequency results in phase
errors .
WIDTH VALUE is used to set the amplitude parameter (for example 3 dB) that
de nes the start and stop points for a bandwidth search. The
bandwidth search feature analyzes a bandpass or band reject
trace and calculates the center point, bandwidth, and Q (quality
factor) for the speci ed bandwidth. Bandwidth units are the
units of the current format.
WIDTHS on OFF turns on the bandwidth search feature and calculates the
center stimulus value , bandwidth, and Q of a bandpass or band
reject shape on the trace . The amplitude value that de nes the
passband or rejectband is set using the
WIDTH VALUE softkey .
F our markers are turned on, and each has a dedicated use .
Marker 1 is a starting point from which the search is begun.
Marker 2 goes to the bandwidth center point. Marker 3 goes to
the bandwidth cuto point on the left, and Marker 4 to the
cuto point on the right.
9-46 K ey De nitions

If a delta marker or xed marker is on, it is used as the
reference point from which the bandwidth amplitude is
measured. F or example , if marker 1 is the delta marker and is
set at the passband maximum, and the width value is set to
3 dB , the bandwidth search nds the bandwidth cuto points
3 dB below the maximum and calculates the 3 dB bandwidth
and Q.
If marker 2 (the dedicated bandwidth center point marker) is
the delta reference marker , the search nds the points 3 dB
down from the center .
If no delta reference marker is set, the bandwidth values are
absolute values .
WINDOW (Option 010 only) is used to specify the parameters of the
window in the transform menu.
allowed. This minimizes the sidelobes and provides the greatest
dynamic range .
WINDOW: MINIMUM
(Option 010 only) is used to set the window of a time domain
measurement to the minimum value . Provides essentially no
window .
measurement to the normal value . Usually the most useful
because it reduces the sidelobes of the measurement somewhat.
is used to terminate basic units: dB , dBm, Hz, dB/GHz, degrees ,
or seconds . It may also be used to terminate unitless entries
such as averaging factor .
XMIT CNTRL [ ]
toggles the PLOTTER/PRINTER serial port data transmit
control mode between the Xon-Xo protocol handshake and
the DTR-DSR (data terminal ready-data set ready) hardwire
handshake .
Y: REFL converts re ection data to its equivalent admittance values .
Y: TRANS
converts transmission data to its equivalent admittance values .
converts re ection data to its equivalent impedance values .
NNNNNNNNNNNNNNNNNNNNNNNNNN
Z: TRANS
converts transmission data to its equivalent impedance values .
K ey De nitions 9-47

Cross Reference of K ey Function to Programming Command
The following table lists the front-panel keys and softkeys alphabetically . The \Command"
column identi es the command that is similar to the front-panel or softkey function. Softkeys
that do not have corresponding programming commands are not included in this section.
T able 9-1. Cross Reference of K ey Function to Programming Command
K ey Name Command
Step Up UP
Step Down DOWN
MODE OFF Delta Marker Mode O DELO
REF = 1 Delta Reference = Marker 1 DELR1
REF = 2 Delta Reference = Marker 2 DELR2
REF = 3 Delta Reference = Marker 3 DELR3
REF = 4 Delta Reference = Marker 4 DELR4
REF = 5 Delta Reference = Marker 5 DELR5
REF = FIXED MKR Delta Reference = Delta Fixed
NNNNNNNNNNN
1/S Inverted S-P arameters CONVIDS
A Measure Channel A MEASA
A/B Ratio of A to B AB
A/R Ratio of A to R AR
ADAPTER:COAX A dapter:Coax ADPTCO AX
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ADAPTER:WAVEGUIDE A dapter:W aveguide ADPTW A VE
ADAPTER DELAY A dapter Delay AD AP1
ADD A dd SADD
ALL SEGS SWEEP All Segments Sweep ASEG
ALTERNATE A and B Alternate A and B AL T AB
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AMPLITUDE OFFSET Amplitude O set LIMIAMPO
ANALOG BUS ON off Analog Bus On ANAB
ANALOG IN Aux Input Analog In ANAI
ARBITRARY IMPEDANCE Arbitrary Impedance STDT ARBI
ASSERT SRQ Service Request ASSS
9-48 K ey De nitions

T able 9-1.
Cross Reference of K ey Function to Programming Command (continued)
K ey Name Command
AUTO FEED ON off (Plotter) Plotter A uto F eed On PL TTRA UTFON
AUTO FEED on OFF (Plotter) Plotter A uto F eed O PL TTRA UTFOFF
AUTO FEED ON off (Printer) Printer A uto F eed On PRNTRA UTFON
AUTO FEED on OFF (Printer) Printer A uto F eed O PRNTRA UTOFF
AUTO SCALE A uto Scale A UTO
AVERAGING FACTOR A veraging F actor A VERF A CT
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AVERAGING ON off A veraging On A VERON
AVERAGING on OFF A veraging O A VEROFF
AVERAGING RESTART A veraging Restart A VERREST
A verage MENU A VG
B Measure Channel B MEASB
B/R Ratio of B to R BR
BACKGROUND INTENSITY Background Intensity B A CI
BANDPASS Bandpass B ANDP ASS
BEEP DONE ON off Beep Done On BEEPDONEON
BEEP DONE on OFF Beep Done O BEEPDONEOFF
BEEP FAIL ON off Beep F ail On BEEPF AILON
BEEP FAIL on OFF Beep F ail O BEEPF AILOFF
BEEP WARN ON off Beep W arn On BEEPW ARNON
BEEP WARN on OFF Beep W arn O BEEPW ARNOFF
BLANK DISPLAY Blank Display On BLADON
NNNNNNNN
C0 C0 T erm C0
C1 C1 T erm C1
C2 C2 T erm C2
C3 C3 T erm C3
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL FACTOR Calibration F actor CALFCALF
CAL FACTOR SENSOR A Calibration F actor Sensor A CALFSENA
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL FACTOR SENSOR B Calibration F actor Sensor B CALFSENB
K ey De nitions 9-49

T able 9-1.
Cross Reference of K ey Function to Programming Command (continued)
K ey Name Command
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CENTER Center , list freq subsweep CENT
CHAN 1 Channel 1 A ctive CHANT
CH1 DATA [ ] Channel 1 Data (Color) PCOLD A T A1
CH1 DATA LIMIT LN Channel 1 Data/Limit Line COLOCH1D
CH1 MEM Channel 1 Memory COLOCH1M
CH1 MEM [ ] Channel 1 Memory (Color) PCOLMEMO1
CHAN 2 Channel 2 A ctive CH2
CH2 DATA [ ] Channel 2 Data (Color) PCOLD A T A2
CH2 DATA LIMIT LN Channel 2 Data/Limit Line COLOCH2D
CH2 MEM [ ] Channel 2 Memory (Color) PCOLMEMO2
CH2 MEM REF LINE Channel 2 Memory Reference Line COLOCH2M
CH PWR [COUPLED] Channel P ower Coupled CHANPCPLD
CH PWR [UNCOUPLED] Channel P ower Uncoupled CHANPUNCPLD
CHOP A and B Chop A and B CHOP AB
CLASS DONE Class Done CLAD
CLEAR BIT Clear Bit CLEABIT
CLEAR LIST Clear List CLEAL
CLEAR SEQUENCE Clear Sequence CLEASEn
COAX Coax CO AX
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
COAXIAL DELAY Coaxial Delay CO AD
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CONTINUE SEQUENCE Continue Sequence CONS
1 CALK35MM selects the HP 85053C cal kit for the HP 8752C/53D/53E.
9-50 K ey De nitions
· · COLOR Color COLOR

T able 9-1.
Cross Reference of K ey Function to Programming Command (continued)
K ey Name Command
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CONVERSION [OFF] Conversion O CONV OFF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CORRECTION ON off Correction On CORRON
CORRECTION on OFF Correction O CORROFF
COUPLED CH ON off Coupled Channel On COUCON
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
COUPLED CH on OFF Coupled Channel O COUCOFF
CW FREQ CW Frequency CWFREQ
CW TIME CW Time CWTIME
D2/D1 to D2 ON off Ratio D2 to D1 On D1DIVD2ON
D2/D1 to D2 on OFF Ratio D2 to D1 O D1DIVD2OFF
DATA and MEMORY Data and Memory DISPD A TM
DATA ARRAY ON off Data Array On EXTMD A T A ON
DATA ARRAY on OFF Data Array O EXTMD A T A OFF
DATA/MEM Ratio Data to Memory DISPDDM
DATA - MEM Data Minus Memory DISPDMM
DATA ! MEMORY Data to Memory D A TI
DATA ONLY ON off Data Only On EXTMD A TOON
DATA ONLY on OFF Data Only O EXTMD A TOOFF
DECR LOOP COUNTER Decrement Loop Counter DECRLOOC
DEFAULT COLORS Default Colors DEFC
DEFAULT PLOT SETUP Default Plot Setup DFL T
DEFAULT PRINT SETUP Default Print Setup DEFLPRINT
DEFINE STANDARD De ne Standard DEFS
NNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DELETE ALL FILES Delete All Files CLEARALL
DELTA LIMITS Delta Limits LIMD
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DIRECTORY SIZE Directory Size DIRS
DISK UNIT NUMBER Disk Unit Number DISCUNIT
DISP MKRS ON off Display Markers On DISM
DISP MKRS on OFF Display Markers O DISM
K ey De nitions 9-51

T able 9-1.
Cross Reference of K ey Function to Programming Command (continued)
K ey Name Command
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DO SEQUENCE Do Sequence DOSEn
NNNNNNNNNNNNNN
DONE (Segment) Done SDON
DONE 1-PORT CAL Done 1-P ort Calibration SA V1
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DONE 2-PORT CAL Done 2-P ort Calibration SA V2
DONE RESPONSE Done Response RESPDONE
DONE RESP ISOL'N CAL Done Response Isolation Cal RAID
DONE SEQ MODIFY Done Sequence Modify DONM
DONE TRL/LRM Done TRL/LRM SA VT
DOWN CONVERTER Down Converter DCONV
DUAL CH ON off Dual Channel On DU A CON
DUAL CH on OFF Dual Channel O DU A COFF
DUPLICATE SEQUENCE Duplicate Sequence DUPLSEQxSEQy
EACH SWEEP Calibrate Each Sweep PWMCEA CS
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
EDIT LIMIT LINE Edit Limit Line EDITLIML
EDIT LIST Edit List EDITLIST
ELECTRICAL DELAY Electrical Delay ELED
EMIT BEEP Emit Beep EMIB
END SWEEP HIGH PULSE End Sweep High Pulse TTLHPULS
END SWEEP LOW PULSE End Sweep Low Pulse TTLLPULS
Entry O ENTO
EXT TRIG ON POINT External Trigger on P oint EXTTPOIN
EXT TRIG ON SWEEP External Trigger on Sweep EXTTON
EXTENSION INPUT A Extension Input A PORT A
EXTENSION INPUT B Extension Input B PORTB
EXTENSION PORT 1 Extension P ort 1 PORT1
9-52 K ey De nitions

T able 9-1.
Cross Reference of K ey Function to Programming Command (continued)
K ey Name Command
EXTENSION PORT 2 Extension P ort 2 PORT2
EXTENSIONS ON off Extensions On POREON
EXTENSIONS on OFF Extensions O POREOFF
EXTERNAL DISK External Disk EXTD
FILENAME FILE0 File Name File 0 TITF0
FILETITLE FILE0 File Name File 0 TITF0
NNNNNNNNNNNNNNNN
FIXED Fixed Load FIXE
FIXED MKR AUX VALUE Fixed Marker A uxiliary V alue MARKF A UV
FIXED MKR POSITION Fixed Marker P osition DELRFIXM
FIXED MKR STIMULUS Fixed Marker Stimulus MARKFSTI
FIXED MKR VALUE Fixed Marker V alue MARKFV AL
FLAT LINE Flat Line LIMTFL
F ormat MENUFORM
FORMAT ARY ON off F ormat Array On EXTMFORMON
FORMAT ARY on OFF F ormat Array O EXTMFORMOFF
FORMAT EXT DISK F ormat External Disk INIE
FORMAT INT DISK F ormat Internal Disk INID
FORMAT INT MEMORY F ormat Internal Memory INTM
FREQ OFFS ON off Frequency O set On FREQOFFSON
FREQ OFFS on OFF Frequency O set O FREQOFFSOFF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FREQUENCY BLANK Frequency Blank FREO
FULL 2-PORT Full 2-P ort CALIFUL2
FULL PAGE Full P age FULP
FWD ISOL'N ISOL'N STD F orward Isolation FWDI
NNNNNNNNNNNNNNNNNNNNNNNNNNNN
FWD MATCH (Label Class) Label F orward Match LABEFWDM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FWD MATCH (Specify Class) Specify F orward Match SPECFWDM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FWD MATCH THRU F orward Match Thru FWDM
K ey De nitions 9-53
· FREQUENCY Frequency CALFFREQ

T able 9-1.
Cross Reference of K ey Function to Programming Command (continued)
K ey Name Command
FWD TRANS (Label Class) Label F orward Transmission LABEFWDT
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FWD TRANS (Specify Class) Specify F orward Transmission SPECFWDT
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FWD TRANS THRU F orward Transmission Thru FWDT
G+jB MKR G+jB Marker Readout SMIMGB
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
GATE ON off Gate On GA TEOON
GATE on OFF Gate O GA TEOOFF
GATE SHAPE MAXIMUM Gate Shape Maximum GA TSMAXI
GATE SHAPE MINIMUM Gate Shape Minimum GA TSMINI
GATE SHAPE NORMAL Gate Shape Normal GA TSNORM
GATE SHAPE WIDE Gate Shape Wide GA TSWIDE
GOSUB SEQUENCE GOSUB Sequence GOSUBn
GRAPHICS ON off Graphics On EXTMGRAPON
GRAPHICS on OFF Graphics O EXTMGRAPOFF
GRATICULE [ ] Print Color - graticule PCOLGRA T
GRATICULE TEXT Graticule COLOGRA T
HARMONIC OFF Harmonic Mode O HARMOFF
HARMONIC SECOND Measure Second Harmonic HARMSEC
HARMONIC THIRD Measure Third Harmonic HARMTHIR
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
HP-IB DIAG ON off HP-IB Diagnostics On DEBUON
HP-IB DIAG on OFF HP-IB Diagnostics O DEBUOFF
IF BW [ ] IF Bandwidth IFBW
IF LIMIT TEST FAIL If Limit T est F ail IFL TF AIL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
IF LIMIT TEST PASS If Limit T est P ass IFL TP ASS
IF LOOP COUNTER = 0 IF Loop Counter = 0 IFLCEQZE
IF LOOP < > COUNTER 0 IF Loop < > Counter 0 IFLCNEZE
IMAGINARY Imaginary IMA G
INCR LOOP COUNTER Increment Loop Counter INCRLOOC
9-54 K ey De nitions
HOLD Hold HOLD · INTENSITY Intensity INTE

T able 9-1.
Cross Reference of K ey Function to Programming Command (continued)
K ey Name Command
INTERNAL DISK Internal Disk INTD
INTERNAL MEMORY Select Internal Memory INTM
INTERPOL ON off Interpolation On CORION
INTERPOL on OFF Interpolation O CORIOFF
ISOLATION (2-P ort) Isolation ISOL
ISOLATION (One-P ath 2-P ort) Isolation ISOOP
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ISOLATION DONE Isolation Done ISOD
ISOL'N STD Isolation Standard RAIISOL
KIT DONE (MODIFIED) Kit Done KITD
LABEL KIT Label Kit LABK
LABEL STD Label Standard LABS
LEFT LOWER Left lower LEFL
LEFT UPPER Left Upper LEFU
LIMIT LINE ON off Limit Line On LIMILINEON
LIMIT LINE on OFF Limit Line O LIMILINEOFF
LIMIT TEST ON off Limit T est On LIMITESTON
LIMIT TEST on OFF Limit T est O LIMITESTOFF
LIN FREQ Linear Frequency LINFREQ
LIN MAG Linear Magnitude LINM
LIN MKR Linear Marker POLMLIN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LINE TYPE DATA Line Type Data LINTD A T A
LINE TYPE MEMORY Line Type Memory LINTMEMO
LIST FREQ List Frequency LISTFREQ
LIST VALUES List V alues LISV
LN/MATCH 1 Line/Match 1 TRLL1
LN/MATCH 2 Line/Match 1 TRLL2
LO CONTROL ON off LO Control On LOCONTON
LO CONTROL on OFF LO Control O LOCONTOFF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LO SOURCE ADDRESS LO Source A ddress ADDRLSRC
LOAD Load STDTLO AD
LOAD NO OFFSET Load No O set LO AN
LOAD OFFSET Load O set LO A O
LOAD SEQ FROM DISK Load Sequence From Disk LO ADSEQn
K ey De nitions 9-55
LINE/MATCH Line/Match

T able 9-1.
Cross Reference of K ey Function to Programming Command (continued)
K ey Name Command
NNNNNNNNNNNNNNNNNNNNNNNNNN
LOG FREQ Logarithmic Frequency LOGFREQ
LOG MAG Logarithmic Magnitude LOGM
LOG MKR Logarithmic Marker SMIMLOG
LOOP COUNTER Loop Counter LOOC
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LOW PASS IMPULSE Low P ass Impulse LOWPIMPU
LOW PASS STEP Low P ass Step LOWPSTEP
LOWER LIMIT Lower Limit LIML
MANUAL TRG ON POINT Manual Trigger On P oint MANTRIG
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER ! CENTER Marker to Center MARKCENT
MARKER ! CW Marker to CW MARKCW
MARKER ! DELAY Marker to Delay MARKDELA
MARKER ! MIDDLE Marker to Middle MARKMIDD
MARKER ! REFERENCE Marker to Reference MARKREF
MARKER ! SPAN Marker to Span MARKSP AN
MARKER ! START Marker to Start MARKST AR
MARKER ! STIMULUS Marker to Stimulus MARKSTIM
MARKER ! STOP Marker to Stop MARKSTOP
MARKER 1 Marker 1 MARK1
MARKER 2 Marker 2 MARK2
MARKER 3 Marker 3 MARK3
MARKER 4 Marker 4 MARK4
MARKER 5 Marker 5 MARK5
MARKER all OFF All Markers O MARKOFF
Marker Function MENUMRKF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
9-56 K ey De nitions

T able 9-1.
Cross Reference of K ey Function to Programming Command (continued)
K ey Name Command
MAXIMUM FREQUENCY Maximum Frequency MAXF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MEASURE RESTART Measure Restart REST
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MIDDLE VALUE Middle V alue LIMM
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MINIMUM FREQUENCY Minimum Frequency MINF
MKR SEARCH [OFF] Marker Search O SEA OFF
MKR ZERO Marker Zero MARKZERO
MODIFY [ ] Modify Kit MODI1
NETWORK ANALYZER Network Analyzer INSMNET A
NEW SEQ/MODIFY SEQ New Sequence/Modify Sequence NEWSEn
NEXT PAGE Display Next P age of T abular NEXP
NUMBER OF GROUPS Number of Groups NUMG
NUMBER OF POINTS Number of P oints POIN
NUMBER OF READINGS Number of Readings NUMR
OFF O CONOFF
NNNNNNNNNNNNNNNNNNNNNNN
OFFSET O set OFLS
OFFSET DELAY O set Delay OFSD
OFFSET LOADS DONE O set Loads Done OFLD
OFFSET LOSS O set Loss OFSL
OFFSET Z0 O set Impedance OFSZ
OMIT ISOLATION Omit Isolation OMII
ONE-PATH 2-PORT One-P ath 2-P ort CALIONE2
ONE SWEEP Calibrate One Sweep PWMCONES
OP PARMS MKRS etc T abular Listing of Operating OPEP
K ey De nitions 9-57

T able 9-1.
Cross Reference of K ey Function to Programming Command (continued)
K ey Name Command
P MTR/HPIB TO TITLE P ower Meter HPIB to Title PMTRTTIT
PARALL IN BIT NUMBER P arallel in Bit Number P ARAIN
PARALL IN IF BIT H P arallel in IF Bit H IFBIHIGH
PARALL IN IF BIT L P arallel in IF Bit L IFBILOW
PARALLEL [COPY] Set parallel port to copy mode P ARALCPY
PARALLEL [GPIO] Set parallel port to GPIO mode P ARALGPIO
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PARALLEL OUT ALL P arallel Out All P ARA OUT
PAUSE P ause P A US
PAUSE TO SELECT P ause to Select PTOS
PEN NUM DATA P en Number Data PENND A T A
PEN NUM GRATICULE P en Number Graticule PENNGRA T
PEN NUM MARKER P en Number Marker PENNMARK
PEN NUM MEMORY P en Number Memory PENNMEMO
PEN NUM TEXT P en Number T ext PENNTEXT
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PHASE OFFSET Phase O set PHA O
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PLOT DATA ON off Plot Data On PD A T A ON
PLOT DATA on OFF Plot Data O PD A T A OFF
PLOT GRAT ON off Plot Graticule On PGRA TON
PLOT GRAT on OFF Plot Graticule O PGRA TOFF
PLOT MEM ON off Plot Memory On PMEMON
PLOT MEM on OFF Plot Memory O PMEMOFF
PLOT MKR ON off Plot Marker ON PMKRON
PLOT MKR on OFF Plot Marker O PMKROFF
PLOTNAME PLOTFILE Plot name Plot le TITP
PLOT SPEED [FAST] Plot Speed F ast PLOSF AST
PLOT SPEED [SLOW] Plot Speed Slow PLOSSLOW
PLOT TEXT ON off Plot T ext On PTEXTON
9-58 K ey De nitions
PHASE Phase PHAS · PLOT Plot PLOT

T able 9-1.
Cross Reference of K ey Function to Programming Command (continued)
K ey Name Command
PLOT TEXT on OFF Plot T ext O PTEXTOFF
PLOTTER BAUD RATE Plotter Baud Rate PL TTRB A UD
PLOTTER FORM FEED Plotter F orm F eed PL TTRFORF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PLTR TYPE [PLOTTER] Plot to a Plotter PL TTYPPL TR
PLTR TYPE [HPGL PRT] Plot to a HP-GL/2 Compatible PL TTYPHPGL
POLAR P olar POLA
PORT PWR [COUPLED] P ort P ower Coupled PORTPCPLD
PORT PWR [UNCOUPLED] P ort P ower Uncoupled PORTPUNCPLD
POWER P ower POWE
POWER MTR P ower Meter POWM
POWER RANGES P ower Ranges PWRR
POWER SWEEP P ower Sweep POWS
F actory Preset RST
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PREVIOUS PAGE Previous P age PREP
PRINT ALL COLOR Print Entire List - Color PRINT ALL
PRINT ALL MONOCHROME Print Entire List - Monochrome PRINT ALL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PRINT COLOR Print Color PRINALL
K ey De nitions 9-59
Printer ·

T able 9-1.
Cross Reference of K ey Function to Programming Command (continued)
K ey Name Command
PRINT MONOCHROME Print Monochrome PRINALL
PRINT SEQUENCE Print Sequence PRINSEQn
PRINTER BAUD RATE Printer Baud Rate PRNTRB A UD
PRINTER FORM FEED Printer F orm F eed PRNTRFORF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PRNTR TYPE [DESKJET] DeskJet Printer PRNTYPDJ
PRNTR TYPE [EPSON-P2] EPSON ESC/P2 Printer Central PRNTYPEP
PRNTR TYPE [LASERJET] LaserJet Printer PRNTYPLJ
PRNTR TYPE [PAINTJET] P aintJet Printer PRNTYPP J
PRNTR TYPE [THINKJET] ThinkJet Printer PRNTYPT J
PWR LOSS ON off P ower Loss On PWRLOSSON
PWR LOSS on OFF P ower Loss O PWRLOSSOFF
PWR RANGE AUTO man P ower Range A uto PWRRP A UTO
PWR RANGE auto MAN P ower Range Man PWRRPMAN
PWRMTR CAL [ ] P ower Meter Calibration CALPOW
PWRMTR CAL [OFF] P ower Meter Calibration O PWRMMCALOFF
R Measure Channel R MEASR
R+jX MKR R+jX Marker Readout SMIMRX
RANGE 0 -15 TO +10 P ower Range 0 PRAN0
RANGE 1 -25 TO 0 P ower Range 1 PRAN1
RANGE 2 -35 TO -10 P ower Range 2 PRAN2
RANGE 3 -45 TO -20 P ower Range 3 PRAN3
RANGE 4 -55 TO -30 P ower Range 4 PRAN4
RANGE 5 -65 TO -40 P ower Range 5 PRAN5
RANGE 6 -75 TO -50 P ower Range 6 PRAN6
RANGE 7 -85 TO -60 P ower Range 7 PRAN7
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RAW ARRAY ON oFF Raw Array On EXTMRA WON
RAW ARRAY on OFF Raw Array O EXTMRA WOFF
RAW OFFSET ON Off Raw O set RA WOFFON
RAW OFFSET On OFF Raw O set RA WOFFSOFF
Re/Im MKR Real/Imaginary Markers POLMRI
9-60 K ey De nitions
Language

T able 9-1.
Cross Reference of K ey Function to Programming Command (continued)
K ey Name Command
READ FILE TITLES Read File Titles REFT
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RECALL COLORS Recall Colors RECO
RECALL REG1 Recall Register 1 RECA1
RECALL REG2 Recall Register 2 RECA2
RECALL REG3 Recall Register 3 RECA3
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RECALL REG4 Recall Register 4 RECA4
RECALL REG5 Recall Register 5 RECA5
RECALL REG6 Recall Register 6 RECA6
RECALL REG7 Recall Register 7 RECA7
RECALL STATE Recall State RECA
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RECEIVER CAL Receiver Calibration REIC
REFERENCE POSITION Reference P osition REFP
REFERENCE VALUE Reference V alue REFV
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
REFLECTION Re ection REFOP
REMOVE ADAPTER Remove A dapter MODS
RESET COLOR Reset Color RSCO
RESPONSE (Calibrate) Response CALIRESP
RESPONSE (Label Class) Response LABERESP
RESPONSE (Specify Class) Response SPECRESP
RESPONSE & ISOL'N Response and Isolation CALIRAI
RESPONSE & ISOL'N Response and Isolation LABERESI
RESPONSE & ISOL'N Response and Isolation SPECRESI
RESTORE DISPLAY Restore Display RESD
K ey De nitions 9-61

T able 9-1.
Cross Reference of K ey Function to Programming Command (continued)
K ey Name Command
RESUME CAL SEQUENCE Resume Calibration Sequence RESC
REV ISOL'N ISOL'N STD Reverse Isolation REVI
REV MATCH (Label Class) Label Reverse Match LABEREVM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
REV MATCH (Specify Class) Specify Reverse Match SPECREVM
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
REV MATCH THRU Reverse Match Thru REVM
REV TRANS (Label Class) Label Reverse Transmission LABEREVT
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
REV TRANS (Specify Class) Specify Reverse Transmission SPECREVT
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
REV TRANS THRU Reverse Transmission Thru REVT
RF > LO RF Greater Than LO RFGTLO
RF < LO RF Less Than LO RFL TLO
RIGHT LOWER Right Lower RIGL
RIGHT UPPER Right Upper RIGU
S11 1-PORT S
NNNNNNNNNNNNNN
S11A S
A Re ected F orward Match LABES11A
S11A S
A Re ected F orward Match SPECS11A
S11B S
B Line F orward Match LABES11B
S11B S
B Line F orward Match SPECS11B
S11C S
C Line F orward Transmission LABES11C
S11C S
C Line F orward Transmission SPECS11C
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
S11 REFL OPEN S
Re ect Short TRLR1
S22 1-PORT S
NNNNNNNNNNNNNN
S22A S
A Re ected Reverse Match LABES22A
S22A S
A Re ected Reverse Match SPECS22A
9-62 K ey De nitions

T able 9-1.
Cross Reference of K ey Function to Programming Command (continued)
K ey Name Command
S22B S
B Line Reverse Transmission LABES22B
S22B S
B Line Reverse Transmission SPECS22B
S22C S
C Line Reverse Transmission LABES22C
NNNNNNNNNNNNN
S22C S
C Line Reverse Transmission SPECS22C
S22 REFL OPEN S
Re ect Short TRLR2
SAMPLR COR ON off Sampler Correction On SAMCON
SAMPLR COR on OFF Sampler Correction O SAMCOFF
SAVE COLORS Save Colors SV CO
SAVE USER KIT Save User Kit SA VEUSEK
SAVE USING ASCII Save ASCII F ormat SA VU ASCI
SAVE USING BINARY Save Using Binary SA VUBINA
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SCALE PLOT [FULL] Scale Plot Full SCAPFULL
SCALE PLOT [GRAT] Scale Plot Graticule SCAPGRA T
Scale Reference MENUSCAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEARCH LEFT Search Left SEAL
SEARCH RIGHT Search Right SEAR
SECOND Second Harmonic HARMSEC
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQUENCE 1 SEQ1 Select Sequence 1 SEQ1
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQUENCE 2 SEQ2 Select Sequence 2 SEQ2
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQUENCE 3 SEQ3 Select Sequence 3 SEQ3
K ey De nitions 9-63

T able 9-1.
Cross Reference of K ey Function to Programming Command (continued)
K ey Name Command
SEQUENCE 4 SEQ4 Select Sequence 4 SEQ4
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQUENCE 5 SEQ5 Select Sequence 5 SEQ5
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQUENCE 6 SEQ6 Select Sequence 6 SEQ6
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQUENCE 1 SEQ1 Select Sequence 1 to Title TITSEQ1
SEQUENCE 2 SEQ2 Select Sequence 2 to Title TITSEQ2
SEQUENCE 3 SEQ3 Select Sequence 3 to Title TITSEQ3
SEQUENCE 4 SEQ4 Select Sequence 4 to Title TITSEQ4
SEQUENCE 5 SEQ5 Select Sequence 5 to Title TITSEQ5
SEQUENCE 6 SEQ6 Select Sequence 6 to Title TITSEQ6
SET BIT Set Bit SETBIT
SET DATE Set Date SETD A TE
SET FREQ LOW PASS Set Frequency Low P ass SETF
SET TIME Set Time SETTIME
SET Z0 Set Impedance SETZ
SHOW MENUS Show Menus SHOM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SINGLE POINT Single P oint LIMTSP
SINGLE SEG SWEEP Single Segment Sweep SSEG
NNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SLOPE ON off Slope On SLOPON
SLOPE on OFF Slope On SLOPOFF
SLOPING LINE Sloping Line LIMTSL
SMITH CHART Smith Chart SMIC
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SMOOTHING APERTURE Smoothing Aperture SMOO APER
SMOOTHING ON off Smoothing On SMOOON
SMOOTHING on OFF Smoothing O SMOOOFF
9-64 K ey De nitions
SINGLE Single SING · SLIDING Sliding SLIL · SLOPE Slope SLOPE

T able 9-1.
Cross Reference of K ey Function to Programming Command (continued)
K ey Name Command
SOURCE PWR ON off Source P ower On SOUPON
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SOURCE PWR on OFF Source P ower O SOUPOFF
NNNNNNNNNNNNNN
SPAN Span SP AN
SPECIFY GATE Specify Gate SPEG
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SPLIT DISP ON off Split Display On SPLDON
SPLIT DISP on OFF Split Display O SPLDOFF
SPUR AVOID ON Off Spur A voidance On SM8ON
SPUR AVOID On OFF Spur A voidance O SM8OFF
Start LOFST AR
START Start ST AR
STATS ON off Statistics On MEAST A TON
STATS on OFF Statistics O MEAST A TOFF
STD DONE (DEFINED) Standard Done STDD
IMPEDANCE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
STEP SIZE Step Size STPSIZE
STIMULUS VALUE Stimulus V alue LIMS
STIMULUS OFFSET Stimulus O set LIMISTIO
NNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
STORE SEQ TO DISK Store Sequence to Disk STORSEQn
SWEEP Sweep Mode LOFSWE
SWEEP TIME [ ] Sweep Time SWET
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SWEEP TIME [AUTO] Sweep Time SWEA
SWR SWR SWR
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SYSTEM CONTROLLER System Controller
TAKE CAL SWEEP T ake Calibration Sweep T AKCS
TAKE RCVR CAL SWEEP T ake Receiver Calibration Sweep T AKRS
K ey De nitions 9-65

T able 9-1.
Cross Reference of K ey Function to Programming Command (continued)
K ey Name Command
TALKER/LISTENER T alker/Listener T ALKLIST
TARGET T arget SEA T ARG
TERMINAL IMPEDANCE T erminal Impedance TERI
TESTPORT (1) 2 T estport 1 TSTPP1
TESTPORT 1 (2) T estport 2 TSTPP2
TESTSET I/O FWD T estset I/O F orward TSTIOFWD
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TESTSET I/O REV T estset I/O Reverse TSTIOREV
TESTSET SW XXXX T estset Switching XXXX TSSWI
NNNNNNNNNNNNNN
TEXT T ext COLOTEXT
TEXT [ ] Print Color - T ext PCOL TEXT
THRU THRU Thru Thru TRL T
TIME STAMP ON off Time Stamp On TIMEST AMON
TIME STAMP on OFF Time Stamp O TIMEST AMOFF
NNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TITLE FILE1 Title File 1 TITF1
TITLE FILE2 Title File 2 TITF2
TITLE FILE3 Title File 3 TITF3
TITLE FILE4 Title File 4 TITF4
TITLE FILE5 Title File 5 TITF5
TITLE SEQUENCE Title Sequence TITSQ
TITLE TO MEMORY Title to Memory TITTMEM
TITLE TO P MTR/HPIB Title to P ower Meter/HPIB TITTPMTR
TITLE TO PERIPHERAL Title to HP-IB P eripheral TITTPERI
TITLE TO PRNTR/HPIB Title to HP-IB Printer TITTPRIN
TRACKING ON off Tracking On TRA CKON
TRACKING on OFF Tracking O TRA CKOFF
TRANS DONE Transmission Done TRAD
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TRANSFORM ON off Transform On TIMDTRANON
TRANSFORM on OFF Transform O TIMDTRANOFF
9-66 K ey De nitions
TINT Tint TINT · TITLE Title TITL

T able 9-1.
Cross Reference of K ey Function to Programming Command (continued)
K ey Name Command
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TRL*/LRM* 2-PORT Thru, Re ect, Line/Line , Re ect,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TRL/LRM OPTION Thru, Re ect, Line/Line , Re ect,
TRL LINE OR MATCH TRL Line or Match SPECTRLL
TRL LINE OR MATCH TRL Line or Match LABETRLL
TRL THRU TRL Thru SPECTRL T
TRL THRU TRL Thru LABETRL T
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TRL REFLECT TRL Re ect SPECTRLR
TRL REFLECT TRL Re ect LABETRLR
TTL OUT HIGH TTL Out High TTLOH
TTL OUT LOW TTL Out Low TTLOL
TUNED RECEIVER Tuned Receiver INSMTUNR
NNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
UP CONVERTER Up Converter UCONV
UPPER LIMIT Upper Limit LIMU
USE MEMORY ON off Use Memory On WINDUSEMON
USE MEMORY on OFF Use Memory O WINDUSEMOFF
USE PASS CONTROL Use P ass Control USEP ASC
USE SENSOR (A) / B Use Sensor A ENSA
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
USE SENSOR A / (B) Use Sensor B ENSB
VELOCITY FACTOR V elocity F actor VELOF A CT
VIEW MEASURE View Measure VIEM
VOLUME NUMBER V olume Number DISCV OLU
WAIT x W ait x Seconds SEQW AIT
K ey De nitions 9-67

T able 9-1.
Cross Reference of K ey Function to Programming Command (continued)
K ey Name Command
WAVEGUIDE W aveguide W A VE
WAVEGUIDE DELAY W aveguide Delay W A VD
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
WIDTH VALUE Width V alue WIDV
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
WIDTHS ON off Widths On WIDTON
WIDTHS on OFF Widths O WIDTOFF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
XMIT CNTRL [Xon-Xoff] Transmit Control (printer) PRNHNDSHKXON
XMIT CNTRL [DTR-DSR] Transmit Control (printer) PRNHNDSHKDTR
XMIT CNTRL [Xon-Xoff] Transmit Control (plotter) PL THNDSHKXON
XMIT CNTRL [DTR-DSR] Transmit Control (plotter) PL THNDSHKDTR
Y: REFL Y : Re ection CONVYREF
Y: TRANS Y : Transmission CONVYTRA
YELLOW Y ellow YELLOW
Z: REFL Z: Re ection CONVZREF
Z: TRANS Z: Transmission CONVZTRA
9-68 K ey De nitions
WHITE White WHITE · WINDOW Window WINDOW

The following table lists the softkey functions alphabetically , and the corresponding front-panel
access key . This table is useful in determining which front-panel key leads to a speci c softkey .
K ey De nitions 9-69
Softkey Locations

T able 9-2. Softkey Locations
A ccess K ey
MODE MENU
MARKER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MODE OFF
MARKER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
REF = 1
MARKER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
REF = 2
MARKER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
REF = 3
MARKER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
REF = 4
MARKER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
REF = 5
MARKER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
REF = FIXED MKR
MARKER
NNNNNNNNNNN
1/S
MEAS
NNNNN
A
MEAS
NNNNNNNNNNN
A/B
MEAS
NNNNNNNNNNN
A/R
MEAS
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ACTIVE ENTRY
DISPLA Y
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ACTIVE MRK MAGNITUDE
DISPLA Y
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ADAPTER:COAX
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ADAPTER:WAVEGUIDE
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ADAPTER DELAY
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ADAPTER REMOVAL
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ADDRESS: 8753
LOCAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ADDRESS: CONTROLLER
LOCAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ADDRESS: DISK
LOCAL
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ADDRESS: DISK
SA VE/RECALL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ADDRESS: P MTR/HPIB
LOCAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ADJUST DISPLAY
DISPLA Y
NNNNNNNNNNNNNNNNNNNNNNN
ALL OFF
MARKER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ALL SEGS SWEEP
MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ALTERNATE A and B
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AMPLITUDE
SYSTEM
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AMPLITUDE OFFSET
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ANALOG IN Aux Input
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ARBITRARY IMPEDANCE
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ASSERT SRQ
SEQ
9-70 K ey De nitions

T able 9-2. Softkey Locations (continued)
A ccess K ey
AUTO FEED on OFF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AUTO SCALE
SCALE REF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AVERAGING FACTOR
A VG
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AVERAGING on OFF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AVERAGING RESTART
A VG
NNNNN
B
MEAS
N
NNNNNNNNNN
B/R
MEAS
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
BACKGROUND INTENSITY
DISPLA Y
NNNNNNNNNNNNNNNNNNNNNNNNNN
BANDPASS
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
BEEP DONE ON off
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
BEEP FAIL on OFF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
BEEP WARN on OFF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
BLANK DISPLAY
DISPLA Y
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
BRIGHTNESS
DISPLA Y
NNNNNNNN
C0
CAL
NNNNNNNN
C1
CAL
NNNNNNNN
C2
CAL
NNNNNNNN
C3
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL FACTOR
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL FACTOR SENSOR A
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL FACTOR SENSOR B
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL KIT [ ]
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL KIT: 2.92*
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
K ey De nitions 9-71

T able 9-2. Softkey Locations (continued)
A ccess K ey
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL KIT: N 50
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL KIT: N 75
CAL
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL KIT: USER KIT
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL ZO: LINE ZO
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CAL ZO: SYSTEM ZO
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CALIBRATE MENU
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CALIBRATE: NONE
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CH1 DATA [ ]
COPY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CH1 DATA LIMIT LN
DISPLA Y
NNNNNNNNNNNNNNNNNNNNNNN
CH1 MEM
DISPLA Y
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CH1 MEM [ ]
COPY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CH2 DATA [ ]
COPY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CH2 DATA LIMIT LN
DISPLA Y
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CH2 MEM [ ]
COPY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CH2 MEM REF LINE
DISPLA Y
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CH PWR [COUPLED]
MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CH PWR [UNCOUPLED]
MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CHOP A and B
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CLEAR BIT
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CLEAR LIST
MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CLEAR SEQUENCE
SEQ
NNNNNNNNNNNNNN
COAX
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
COAXIAL DELAY
SCALE REF
NNNNNNNNNNNNNNNNN
COLOR
DISPLA Y
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CONFIGURE
SYSTEM
N
CONFIGURE EXTERNAL DISK
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SA VE/RECALL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CONTINUE SEQUENCE
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CONTINUOUS
MENU
9-72 K ey De nitions

T able 9-2. Softkey Locations (continued)
A ccess K ey
CONVERSION [ ]
MEAS
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CORRECTION on OFF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
COUPLED CH on OFF
NNNNNNNNNNNNNNNNNNNNNNN
CW FREQ
MENU
NNNNNNNNNNNNNNNNNNNNNNN
CW TIME
MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
D2/D1 to D2 on OFF
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DATA and MEMORY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DATA ARRAY on OFF
NNNNNNNNNNNNNNNNNNNNNNNNNN
DATA/MEM
DISPLA Y
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DATA - MEM
DISPLA Y
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DATA ! MEMORY
DISPLA Y
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DATA ONLY on OFF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DECISION MAKING
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DECR LOOP COUNTER
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DEFAULT COLORS
DISPLA Y
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DEFAULT PLOT SETUP
COPY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DEFAULT PRINT SETUP
COPY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DEFINE DISK-SAVE
SA VE/RECALL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DEFINE PLOT
COPY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DEFINE PRINT
COPY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DEFINE STANDARD
CAL
NNNNNNNNNNNNNNNNN
DELAY
F ORMA T
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DELAY/THRU
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DELETE ALL FILES
SA VE/RECALL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DELETE FILE
SA VE/RECALL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DELTA LIMITS
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DEMOD: AMPLITUDE
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DEMOD: OFF
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DEMOD: PHASE
SYSTEM
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DIRECTORY SIZE (LIF)
SA VE/RECALL
K ey De nitions 9-73

T able 9-2. Softkey Locations (continued)
A ccess K ey
DISK UNIT NUMBER
LOCAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DISK UNIT NUMBER
SA VE/RECALL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DISPLAY: DATA
DISPLA Y
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DISP MKRS ON off
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DO BOTH FWD + REV
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DO SEQUENCE
SEQ
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DONE 1-PORT CAL
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DONE 2-PORT CAL
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DONE RESPONSE
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DONE RESP ISOL'N CAL
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DONE SEQ MODIFY
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DONE TRL/LRM
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DOWN CONVERTER
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DUAL CH on OFF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DUMP GRAPH on OFF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DUPLICATE SEQUENCE
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
EACH SWEEP
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
EDIT LIMIT LINE
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
EDIT LIST
MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ELECTRICAL DELAY
SCALE REF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
EMIT BEEP
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
END OF LABEL
DISPLA Y
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
END SWEEP HIGH PULSE
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
END SWEEP LOW PULSE
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ERASE TITLE
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ERASE TITLE
DISPLA Y
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ERASE TITLE
SA VE/RECALL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
EXT SOURCE AUTO
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
EXT SOURCE MANUAL
SYSTEM
9-74 K ey De nitions

T able 9-2. Softkey Locations (continued)
A ccess K ey
EXT TRIG ON POINT
MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
EXT TRIG ON SWEEP
MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
EXTENSION INPUT A
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
EXTENSION INPUT B
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
EXTENSION PORT 1
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
EXTENSION PORT 2
CAL
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
EXTENSIONS on OFF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
EXTERNAL DISK
SA VE/RECALL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FILETITLE FILE0
SA VE/RECALL
NNNNNNNNNNNNNNNNNNNNNNNNNN
FILENAME
SA VE/RECALL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FILE UTILITES
SA VE/RECALL
NNNNNNNNNNNNNNNNN
FIXED
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FIXED MKR AUX VALUE
MARKER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FIXED MKR POSITION
MARKER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FIXED MKR STIMULUS
MARKER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FIXED MKR VALUE
MARKER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FLAT LINE
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FORM FEED
DISPLA Y
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FORMAT ARY on OFF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FORMAT DISK
SA VE/RECALL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FORMAT: DOS
SA VE/RECALL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FORMAT: LIF
SA VE/RECALL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FORMAT EXT DISK
SA VE/RECALL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FORMAT INT DISK
SA VE/RECALL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FORMAT INT MEMORY
SA VE/RECALL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FREQ OFFS MENU
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FREQ OFFS on OFF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FREQUENCY
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FREQUENCY BLANK
DISPLA Y
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FREQUENCY: CW
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FREQUENCY: SWEEP
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FULL 2-PORT
CAL
K ey De nitions 9-75

T able 9-2. Softkey Locations (continued)
A ccess K ey
FULL PAGE
COPY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FWD ISOL'N ISOL'N STD
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FWD MATCH
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FWD MATCH THRU
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FWD TRANS
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FWD TRANS THRU
CAL
N
NNNNNNNNNNNNNNNNNNNNNNNNN
MARKER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
GATE: CENTER
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
GATE: SPAN
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
GATE: START
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
GATE: STOP
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
GATE on OFF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
GATE SHAPE
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
GATE SHAPE MAXIMUM
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
GATE SHAPE MINIMUM
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
GATE SHAPE NORMAL
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
GOSUB SEQUENCE
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
GRAPHICS on OFF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
GRATICULE [ ]
COPY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
GRATICULE TEXT
DISPLA Y
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
HARMONIC MEAS
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
HARMONIC OFF
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
HARMONIC SECOND
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
HARMONIC THIRD
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
HELP ADAPT REMOVAL
CAL
NNNNNNNNNNNNNN
HOLD
MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
HP-IB DIAG on off
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
IF BW [ ]
A V G
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
IF LIMIT TEST FAIL
SEQ
N
IF LIMIT TEST PASS
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
IF LOOP COUNTER = 0
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
IF LOOP < > COUNTER 0
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
IMAGINARY
F ORMA T
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
INCR LOOP COUNTER
SEQ
9-76 K ey De nitions

T able 9-2. Softkey Locations (continued)
A ccess K ey
INIT DISK? YES
SA VE/RECALL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
INITIALIZE DISK
SA VE/RECALL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
INPUT PORTS
MEAS
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
INSTRUMENT MODE
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
INTENSITY
DISPLA Y
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
INTERNAL DISK
SA VE/RECALL
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
INTERNAL MEMORY
SA VE/RECALL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
INTERPOL on OFF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ISOLATION
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ISOLATION DONE
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ISOL'N STD
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ISTATE CONTENTS
SA VE/RECALL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
KIT DONE (MODIFIED)
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LABEL CLASS
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LABEL CLASS DONE
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LABEL KIT
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LABEL STD
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LEFT LOWER
COPY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LEFT UPPER
COPY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LIMIT LINE OFFSETS
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LIMIT LINE on OFF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LIMIT MENU
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LIMIT TEST on OFF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LIMIT TEST RESULT
DISPLA Y
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LIMIT TYPE
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNN
LIN FREQ
MENU
NNNNNNNNNNNNNNNNNNNNNNN
LIN MAG
F ORMA T
NNNNNNNNNNNNNNNNNNNNNNN
LIN MKR
MARKER F CTN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LIST FREQ
MENU
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LINE/MATCH
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LINE TYPE DATA
COPY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LINE TYPE MEMORY
COPY
NNNNNNNNNNNNNN
LIST
COPY
K ey De nitions 9-77

T able 9-2. Softkey Locations (continued)
A ccess K ey
LN/MATCH 1
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LN/MATCH 2
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LO CONTROL on OFF
NNNNNNNNNNNNNNNNNNNNNNN
LO MENU
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LO SOURCE ADDRESS
SYSTEM
NNNNNNNNNNNNNN
LOAD
CAL
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LOAD NO OFFSET
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LOAD OFFSET
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LOAD SEQ FROM DISK
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNN
LOG FREQ
MENU
NNNNNNNNNNNNNNNNNNNNNNN
LOG MAG
F ORMA T
NNNNNNNNNNNNNNNNNNNNNNN
LOG MKR
MARKER F CTN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LOOP COUNTER
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LOOP COUNTER
DISPLA Y
NNNNNNNNNNNNNN
LOSS
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LOSS/SENSR LISTS
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LOWER LIMIT
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LOW PASS IMPULSE
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LOW PASS STEP
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MANUAL TRG ON POINT
MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER ! AMP. OFS.
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER ! CENTER
MARKER F CTN
9-78 K ey De nitions

T able 9-2. Softkey Locations (continued)
A ccess K ey
MARKER ! CW
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER ! DELAY
MARKER F CTN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER ! DELAY
SCALE REF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER ! MIDDLE
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER ! REFERENCE
MARKER F CTN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER ! REFERENCE
SCALE REF
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER ! SPAN
MARKER F CTN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER ! START
MARKER F CTN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER ! STIMULUS
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER ! STOP
MARKER F CTN
NNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER 1
MARKER
NNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER 2
MARKER
NNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER 3
MARKER
NNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER 4
MARKER
NNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER 5
MARKER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER all OFF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER MODE MENU
MARKER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKERS: CONTINUOUS
MARKER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKERS: COUPLED
MARKER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKERS: DISCRETE
MARKER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKERS: UNCOUPLED
MARKER
NNNNNNNNNNN
MAX
MARKER F CTN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MAXIMUM FREQUENCY
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MEASURE RESTART
MENU
NNNNNNNNNNNNNNNNNNNN
MEMORY
DISPLA Y
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MIDDLE VALUE
SYSTEM
NNNNNNNNNNN
MIN
MARKER F CTN
NNNNNNNNNNNNNNNNNNNNNNN
MINIMUM
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MINIMUM FREQUENCY
CAL
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MKR SEARCH [ ]
MARKER F CTN
NNNNNNNNNNNNNNNNNNNNNNNNNN
MKR ZERO
MARKER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MODIFY [ ]
CAL
K ey De nitions 9-79

T able 9-2. Softkey Locations (continued)
A ccess K ey
MODIFY COLORS
DISPLA Y
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NETWORK ANALYZER
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NEW SEQ/MODIFY SEQ
SEQ
NNNNNNNNNNNNNNNNNNNNNNN
NEWLINE
DISPLA Y
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NEXT PAGE
COPY
NNNNNNNNNNNNNNNNNNNN
NORMAL
SYSTEM
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NUMBER OF GROUPS
MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NUMBER OF POINTS
MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NUMBER OF READINGS
CAL
NNNNNNNNNNNNNNNNNNNN
OFFSET
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
OFFSET DELAY
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
OFFSET LOADS DONE
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
OFFSET LOSS
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
OFFSET Z0
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
OMIT ISOLATION
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ONE-PATH 2-PORT
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ONE SWEEP
CAL
NNNNNNNNNNNNNN
OPEN
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
OP PARMS (MKRS etc)
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
P MTR/HPIB TO TITLE
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PARALL IN BIT NUMBER
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PARALL IN IF BIT H
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PARALL IN IF BIT L
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNN
PARALLEL
LOCAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PARALLEL [ ]
LOCAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PARALLEL OUT ALL
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PAUSE TO SELECT
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PEN NUM DATA
COPY
9-80 K ey De nitions

T able 9-2. Softkey Locations (continued)
A ccess K ey
PEN NUM GRATICULE
COPY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PEN NUM MARKER
COPY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PEN NUM MEMORY
COPY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PEN NUM TEXT
COPY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PERIPHERAL HPIB ADDR
SEQ
NNNNNNNNNNNNNNNNN
PHASE
F ORMA T
N
NNNNNNNNNNNNNNNN
PHASE
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PHASE OFFSET
SCALE REF
NNNNNNNNNNNNNN
PLOT
COPY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PLOT DATA ON off
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PLOT GRAT ON off
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PLOT MEM ON off
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PLOT MKR ON off
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PLOT SPEED [ ]
COPY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PLOT TEXT ON off
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PLOTTER BAUD RATE
LOCAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PLOTTER FORM FEED
COPY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PLOTTER PORT
LOCAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PLTR PORT: DISK
LOCAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PLTR PORT: HPIB
LOCAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PLTR PORT: PARALLEL
LOCAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PLTR PORT: SERIAL
LOCAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PLTR TYPE [ ]
LOCAL
NNNNNNNNNNNNNNNNN
POLAR
F ORMA T
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
POLAR MKR MENU
MARKER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PORT EXTENSIONS
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PORT PWR [COUPLED]
MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PORT PWR [UNCOUPLED]
MENU
NNNNNNNNNNNNNNNNN
POWER
MENU
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
POWER: FIXED
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
POWER: SWEEP
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
POWER LOSS
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
POWER MTR [ ]
LOCAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
POWER RANGES
MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
POWER SWEEP
MENU
K ey De nitions 9-81

T able 9-2. Softkey Locations (continued)
A ccess K ey
PRESET: FACTORY
PRESET
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PRESET: USER
PRESET
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PREVIOUS PAGE
COPY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PRINT: COLOR
COPY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PRINT COLORS
COPY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PRINT: MONOCHROME
COPY
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PRINT MONOCHROME
COPY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PRINT SEQUENCE
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PRINTER BAUD RATE
LOCAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PRINTER FORM FEED
COPY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PRINTER PORT
LOCAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PRNTR PORT: HPIB
LOCAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PRNTR PORT: PARALLEL
LOCAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PRNTR PORT: SERIAL
LOCAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PRNTR TYPE [ ]
LOCAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PWR LOSS on OFF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PWR RANGE AUTO man
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PWRMTR CAL [ ]
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PWRMTR CAL [OFF]
CAL
NNNNN
R
MEAS
NNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RANGE 0 -15 TO +10
MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RANGE 1 -25 TO 0
MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RANGE 2 -35 TO -10
MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RANGE 3 -45 TO -20
MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RANGE 4 -55 TO -30
MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RANGE 5 -65 TO -40
MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RANGE 6 -75 TO -50
MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RANGE 7 -85 TO -60
MENU
9-82 K ey De nitions

T able 9-2. Softkey Locations (continued)
A ccess K ey
RAW ARRAY on OFF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RAW OFFSET ON Off
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER
NNNNNNNNNNNNNN
REAL
F ORMA T
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RECALL CAL PORT 1
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RECALL CAL PORT 2
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RECALL CAL SETS
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RECALL COLORS
DISPLA Y
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RECALL KEYS MENU
SA VE/RECALL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RECALL KEYS on OFF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RECALL REG1
SA VE/RECALL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RECALL REG2
SA VE/RECALL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RECALL REG3
SA VE/RECALL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RECALL REG4
SA VE/RECALL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RECALL REG5
SA VE/RECALL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RECALL REG6
SA VE/RECALL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RECALL REG7
SA VE/RECALL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RECALL STATE
SA VE/RECALL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RECEIVER CAL
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
REFERENCE POSITION
SCALE REF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
REFERENCE VALUE
SCALE REF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
REFL: FWD S11 (A/R)
MEAS
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
REFL: REV S22 (B/R)
MEAS
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
REFLECT AND LINE
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
REFLECTION
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
REMOVE ADAPTER
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RENAME FILE
SA VE/RECALL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RE-SAVE STATE
SA VE/RECALL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RESET COLOR
DISPLA Y
N
NNNNNNNNNNNNNNNNNNNNNNNNN
RESPONSE
CAL
K ey De nitions 9-83

T able 9-2. Softkey Locations (continued)
A ccess K ey
RESPONSE & ISOL'N
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RESUME CAL SEQUENCE
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
REV ISOL'N ISOL'N STD
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
REV MATCH
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
REV MATCH THRU
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
REV TRANS
CAL
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
REV TRANS THRU
CAL
NNNNNNNNNNNNNNNNNNNNNNN
RF > LO
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNN
RF < LO
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RIGHT LOWER
COPY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RIGHT UPPER
COPY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ROUND SECONDS
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
S PARAMETERS
MEAS
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
S11 1-PORT
CAL
NNNNNNNNNNNNNN
S11A
CAL
NNNNNNNNNNNNNN
S11B
CAL
NNNNNNNNNNNNNN
S11C
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
S11 REFL OPEN
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
S22 1-PORT
CAL
NNNNNNNNNNNNNN
S22A
CAL
NNNNNNNNNNNNNN
S22B
CAL
NNNNNNNNNNNNNN
S22C
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
S22 REFL OPEN
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SAMPLR COR ON off
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SAVE COLORS
DISPLA Y
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SAVE USER KIT
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SAVE USING ASCII
SA VE/RECALL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SAVE USING BINARY
SA VE/RECALL
9-84 K ey De nitions

T able 9-2. Softkey Locations (continued)
A ccess K ey
SCALE/DIV
SCALE REF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SCALE PLOT [ ]
COPY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEARCH LEFT
MARKER F CTN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEARCH RIGHT
MARKER F CTN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEARCH: MAX
MARKER F CTN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEARCH: MIN
MARKER F CTN
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEARCH: OFF
MARKER F CTN
NNNNNNNNNNNNNNNNNNNN
SECOND
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNN
SEGMENT
CAL
NNNNNNNNNNNNNNNNNNNNNNN
SEGMENT
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEGMENT: CENTER
MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEGMENT: SPAN
MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEGMENT: START
MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEGMENT: STOP
MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEL QUAD [ ]
COPY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SELECT DISK
SA VE/RECALL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQUENCE 1 SEQ1
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQUENCE 2 SEQ2
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQUENCE 3 SEQ3
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQUENCE 4 SEQ4
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQUENCE 5 SEQ5
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQUENCE 6 SEQ6
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEQUENCE FILENAMING
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SET ADDRESSES
LOCAL
NNNNNNNNNNNNNNNNNNNNNNN
SET BIT
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SET CLOCK
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNN
SET DAY
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SET FREQ LOW PASS
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNN
SET HOUR
SYSTEM
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SET MINUTES
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SET MONTH
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SET REF: THRU
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SET REF: REFLECT
SYSTEM
K ey De nitions 9-85

T able 9-2. Softkey Locations (continued)
A ccess K ey
SET YEAR
SYSTEM
NNNNNNNNNNNNNNNNNNNN
SET Z0
CAL
NNNNNNNNNNNNNNNNN
SHORT
CAL
NNNNNNNNNNNNNNNNNNNN
SINGLE
MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SINGLE POINT
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SINGLE SEG SWEEP
MENU
N
NNNNNNNNNNNNNNNNNNNNNN
SLIDING
CAL
NNNNNNNNNNNNNNNNN
SLOPE
MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SLOPE on OFF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SLOPING LINE
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SMITH CHART
F ORMA T
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SMITH MKR MENU
MARKER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SMOOTHING APERTURE
A V G
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SMOOTHING on OFF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SOURCE PWR ON off
NNNNNNNNNNNNNN
SPAN
MENU
NNNNNNNNNNNNNN
SPAN
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SPECIAL FUNCTIONS
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SPECIFY CLASS
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SPECIFY GATE
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SPECIFY OFFSET
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SPLIT DISP on OFF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SPUR AVOID On Off
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
STANDARDS DONE
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
STATS on OFF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
STD DONE (MODIFIED)
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
STD OFFSET DONE
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
STD TYPE:
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
STEP SIZE
MENU
N
STIMULUS VALUE
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
STIMULUS OFFSET
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
STORE SEQ TO DISK
SEQ
9-86 K ey De nitions

T able 9-2. Softkey Locations (continued)
A ccess K ey
SWEEP
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SWEEP TIME [ ]
MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SWEEP TYPE MENU
MENU
NNNNNNNNNNN
SWR
F ORMA T
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SYSTEM CONTROLLER
LOCAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TAKE CAL SWEEP
CAL
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TAKE RCVR CAL SWEEP
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TALKER/LISTENER
LOCAL
NNNNNNNNNNNNNNNNNNNN
TARGET
MARKER F CTN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TERMINAL IMPEDANCE
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TEST PORT 1 2
MEAS
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TESTSET I/O FWD
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TESTSET I/O REV
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TESTSET SW XXXX
SYSTEM
NNNNNNNNNNNNNN
TEXT
DISPLA Y
NNNNNNNNNNNNNNNNNNNNNNNNNN
TEXT [ ]
COPY
NNNNNNNNNNNNNN
THRU
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
THRU THRU
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TIME STAMP ON off
NNNNNNNNNNNNNN
TINT
DISPLA Y
NNNNNNNNNNNNNNNNN
TITLE
DISPLA Y
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TITLE SEQUENCE
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TITLE TO MEMORY
SEQ
K ey De nitions 9-87

T able 9-2. Softkey Locations (continued)
A ccess K ey
TITLE TO P MTR/HPIB
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TITLE TO PERIPHERAL
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TITLE TO PRNTR/HPIB
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TRACKING on OFF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TRANS DONE
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TRANS: FWD S21 (B/R)
MEAS
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TRANS: REV S12 (B/R)
MEAS
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TRANSFORM MENU
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TRANSFORM on OFF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TRANSMISSION
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TRIGGER MENU
MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TRIGGER: TRIG OFF
MENU
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TRL*/LRM* 2-PORT
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TRL/LRM OPTION
CAL
NNNNNNNNNNNNNNNNNNNNNNN
TTL I/O
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TTL OUT HIGH
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TTL OUT LOW
SEQ
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TUNED RECEIVER
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
UNCOUPLED
MARKER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
UP CONVERTER
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
UPPER LIMIT
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
USE MEMORY on OFF
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
USE PASS CONTROL
LOCAL
NNNNNNNNNNNNNN
USER
PRESET
NNNNNNNNNNNNNNNNNNNNNNNNNN
USER KIT
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
USE SENSOR A / B
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
VELOCITY FACTOR
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
VIEW MEASURE
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
VOLUME NUMBER
LOCAL
N
VOLUME NUMBER
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SA VE/RECALL
NNNNNNNNNNNNNNNNNNNN
SEQ
NNNNNNNNNNNNNNNNNNNNNNN
WARNING
DISPLA Y
9-88 K ey De nitions

T able 9-2. Softkey Locations (continued)
A ccess K ey
WARNING [ ]
COPY
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
WAVEGUIDE
CAL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
WAVEGUIDE DELAY
SCALE REF
NNNNNNNNNNNNNN
WIDE
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
WIDTH VALUE
MARKER F CTN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
WIDTHS on OFF
N
NNNNNNNNNNNNNNNNNNN
WINDOW
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
WINDOW: MAXIMUM
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
WINDOW: MINIMUM
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
WINDOW: NORMAL
SYSTEM
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
XMIT CNTRL [ ]
LOCAL
NNNNNNNNNNNNNNNNNNNNNNN
Y: REFL
MEAS
NNNNNNNNNNNNNNNNNNNNNNNNNN
Y: TRANS
MEAS
NNNNNNNNNNNNNNNNNNNNNNN
Z: REFL
MEAS
NNNNNNNNNNNNNNNNNNNNNNNNNN
Z: TRANS
MEAS
K ey De nitions 9-89



This chapter contains the following information to help you interpret any error messages that
may be displayed on the analyzer LCD or transmitted by the instrument over HP-IB:
An alphabetical listing of all error messages , including:
An explanation of the message
Suggestions to help solve the problem
A numerical listing of all error messages
Some messages described in this chapter are for information only and do not
indicate an error condition. These messages are not numbered and so they will
not appear in the numerical listing.
Where to Look for More Information
A dditional information about many of the topics discussed in this chapter is located in the
following areas:
Chapter 2, \Making Measurements ," contains step-by-step procedures for making
measurements or using particular functions .
Chapter 4, \Printing, Plotting, and Saving Measurement Results ," contains instructions
for saving to disk or the analyzer internal memory , and printing and plotting displayed
measurements .
Chapter 6, \Application and Operation Concepts ," contains explanatory-style information
about many applications and analyzer operation.
Chapter 9, \Key De nitions ," describes all the front panel keys , softkeys , and their
corresponding HP-IB commands .
Chapter 12, \Preset State and Memory Allocation," contains a discussion of memory
allocation, memory storage , instrument state de nitions , and preset conditions .
Error Messages · Error Messages 10-1

Error Messages in Alphabetical Order
This message is displayed brie y if you have pressed
to abort a copy
operation. If the message is not subsequently replaced by error message
number 25, PRINT ABORTED , the copy device may be hung. Press
more to exit the abort process and verify the status of the copy device . At this
point, the copy device will probably have an error condition which must be
Error correction for the selected calibration class cannot be computed until you
have measured all the necessary standards .
Y ou have sent a read command to the analyzer (such as ENTER 716) without
rst requesting data with an appropriate output command (such as OUTPD A T A).
The analyzer has no data in the output queue to satisfy the request.
Something is restricting the air ow into the analyzer . Check for any debris and
clean or replace the fan lter .
Y ou have used all of the available registers; you can store no more instrument
states even though you may still have sucient memory . There are 31 registers
available , plus the present instrument state .
Y ou must remove the active controller from the bus or the controller must
relinquish the bus before the analyzer can assume the system controller mode .

The citi le you just downloaded over the HP-IB or via disk was not properly
organized. The analyzer is unable to read the \BEGIN" statement.
The citi le you just downloaded over the HP-IB or via disk was not properly
organized. The analyzer is unable to read the \CITIFILE" statement.
The citi le you just downloaded over the HP-IB or via disk was not properly
organized. The analyzer is unable to read the \D A T A" statement.
The citi le you just downloaded over the HP-IB or via disk was not properly
organized. The analyzer is unable to read the \V AR" statement.
Y ou cannot use sweep-to-sweep averaging in single-input measurements .
Sweep-sweep averaging is valid only for ratioed measurements (A/R, B/R, A/B ,
and S-parameters). Y ou can use noise reduction techniques , such as narrower
IF bandwidth, for single input measurements .
Y ou turned on time domain or recalled a calibration that resulted in start and
stop frequencies that are beyond the allowable limits .
The battery protection of the non-volatile CMOS memory has failed. The CMOS
memory has been cleared. Refer to the HP 8753D Network Analyzer Service
Guide for battery replacement instructions . See Chapter 12, \Preset State and
Memory Allocation," for more information about the CMOS memory .

The battery protection of the non-volatile CMOS memory is in danger of failing.
If this occurs , all of the instrument state registers stored in CMOS memory will
be lost. Save these states to a disk and refer to the HP 8753D Network Analyzer
Service Guide for battery replacement instructions . See Chapter 12, \Preset
State and Memory Allocation," for more information about the CMOS memory .
The analyzer did not receive a complete data transmission. This is usually
caused by an interruption of the bus transaction. Clear by pressing the
key or aborting the I/O process at the controller .
The length of the header received by the analyzer did not agree with the size
of the internal array block. Refer to the HP 8753D Network Analyzer
Programmer's Guide for instructions on using analyzer input commands .
Y ou have changed the active channel during a calibration so the calibration in
progress was terminated. Make sure the appropriate channel is active and
restart the calibration.
A calibration set could not be found that matched the current stimulus state or
measurement parameter . Y ou will have to perform a new calibration.
Y ou have attempted to initialize a oppy disk to DOS format on an external
disk drive that does not support writing to all 80 tracks of the double density
and high density disks . The older single-sided disks had only 66 tracks and
some disk drives were limited to accessing that number of tracks . T o format
the disk, either choose another external disk drive or use the analyzer's
internal disk drive .

Y ou have attempted to rename , delete , or otherwise alter the factory preset
state . The factory preset state is permanently stored in CMOS memory and
cannot be altered. If your intent was to create a user preset state , you must
create a new instrument state , save it, and then rename it to \UPRESET".
Refer to Chapter 12, \Preset State and Memory Allocation," for more detailed
instructions .
The disk is being accessed by the analyzer and is found to contain an HFS
(hierarchical le system) or les nested within subdirectories . The analyzer
does not support HFS. Replace the disk medium with a LIF or DOS formatted
disk that does not contain les nested within subdirectories .
Y our sequence transfer to or from a disk could not be completed due to
insucient memory .
Y our target value for the marker search function does not exist on the current
data trace .
Y our current measurement requires di erent power ranges on channel 1 and
channel 2. T o protect the attenuator from undue mechanical wear , test set hold
will be activated.
The \tsH" (test set hold) indicator in the left margin of the display indicates
that the inactive channel has been put in the sweep hold mode .
The printer or plotter is not accepting data. V erify the cable connections , HP-IB
addresses , and otherwise ensure that the copy device is ready .

The analyzer has completed outputting data to the printer or plotter . The
analyzer can now accept another copy command.
When you change the frequency range , sweep type , or number of points ,
error-correction is switched o and the time domain transform is recalculated,
without error-correction. Y ou can either correct the frequency range , sweep
type , or number of points to match the calibration, or perform a new
calibration. Then perform a new time domain transform.
A store operation to the EEPROM was not successful. Y ou must change the
position of the jumper on the A9 CPU assembly . Refer to the \A9 CC Jumper
P osition Procedure" in the \A djustments and Correction Constants" chapter of
the HP 8753D Network Analyzer Service Guide.
Critical parameters in your current instrument state do not match the
parameters for the calibration set, therefore correction has been turned o .
The critical instrument state parameters are sweep type , start frequency ,
frequency span, and number of points .
Correction is not valid for your selected measurement parameter . Either
change the measurement parameters or perform a new calibration.
Y ou can only make a D2/D1 measurement if both channels are on.

Y ou can only make a D2/D1 measurement if both channels have the same
number of points .
A fatal rmware error occurred before instrument preset completed. Call your
local Hewlett-P ackard sales and service oce .
Demodulation was selected when the analyzer was not in CW time mode .
Select demodulation only after putting the analyzer into CW time mode .
The device at the selected address cannot be accessed by the analyzer . V erify
that the device is switched on, and check the HP-IB connection between the
analyzer and the device . Ensure that the device address recognized by the
analyzer matches the HP-IB address set on the device itself.
There is no room left in the directory to add les . Either delete les or get a
new disk.
The disk drive is not responding correctly . Refer to the HP 8753D Network
Analyzer Service Guide for troubleshooting information. If using an external
disk drive , refer to the disk drive operating manual.

The store operation cannot write to a write-protected disk. Slide the
write-protect tab over the write-protect opening in order to write data on the
disk.
Y ou must initialize the disk before it can be used.
The analyzer and the external disk drive aren't communicating properly . Check
the HP-IB connection and then try substituting another disk drive to isolate the
problem instrument.
The disk cannot be accessed by the analyzer . V erify power to the disk drive ,
and check the HP-IB connection between the analyzer and the disk drive .
Ensure that the disk drive address recognized by the analyzer matches the
HP-IB address set on the disk drive itself.
There may be a problem with your disk. Try a new oppy disk. If a new oppy
disk does not eliminate the error , suspect hardware problems .
Cumulative use of the disk is approaching the maximum. Copy les as
necessary using an external controller . If no controller is available , load
instrument states from the old disk and store them to a newly initialized disk
using the save/recall features of the analyzer . Discard the old disk.
Time domain calculations were reset due to a change in the frequency range ,
sweep type , or number of points . P erform a new time domain transform on the
new state .

DOS NAME LIMITED TO 8 CHARS + 3 CHAR EXTENSION
A DOS le name must meet the following criteria:
minimum of 1 character
format is filename.ext
maximum of 8 characters in the lename
maximum of 3 characters in the extension eld (optional)
a dot separates the lename from the extension eld (the dot is not part of
the name on the disk)
A sequence cannot be duplicated to itself.
When modifying calibration kits , you can de ne a maximum of seven standards
for any class .
An external source can only be phase locked and measured in the CW time
sweep mode .
There is a hardware problem with the HP 8625A external source . V erify the
connections between the analyzer and the external source . If the connections
are correct, refer to the source operating manual.
The analyzer is unable to communicate with the external source . Check the
connections and the HP-IB address on the source .

Y ou cannot recall user graphics that had been saved on an earlier model of
analyzer with a monochrome display . These les cannot be used with the
HP 8753D .
The requested le was not found on the current disk medium.
During a resave operation, either the le was not found or the type of le was
not an instrument state le .
The rst character of a disk le title or an internal save register title must be
an alpha character .
Conversion measurements (Z or Y re ection and transmission) are not valid
with Smith chart and SWR formats .
The list information is being processed for a list data output to a copy device
and stored in the copy spool bu er . During this time , the analyzer's resources
are dedicated to this task (which takes less than a few seconds).
Y ou can only make frequency o set measurements in the network analyzer
mode .

The number of points selected for setting the low pass transform frequencies is
too high. Reduce the number of points so that the low pass criteria is met.
The function you requested over HP-IB is not available on the current
instrument.
The function you requested is incompatible with the current instrument state .
Y ou cannot perform sequencing operations while a sequence is being modi ed.
The function you selected only works with disk les .
Y ou can only use the
GOSUB SEQUENCE capability when you are building a
sequence . Attempting to use this softkey at any other time returns an error
message and no action is taken.
Some of the stimulus parameters of the instrument state have been changed,
because you have turned correction on. A calibration set for the current
measurement parameter was found and activated. The instrument state was
updated to match the stimulus parameters of the calibration state .
This message also appears when you have turned on harmonic mode or
frequency o set and the present frequency range cannot be used with one of
these modes .

An HP-IB copy was already in progress when you requested the HP-IB for
another function. T o abort the rst copy , press
, otherwise the HP-IB is
unavailable until the rst copy is completed.
The disk unit or volume number set in the analyzer is not valid. Refer to the
disk drive operating manual.
INIT DISK removes all data from disk
Continuing with the initialize operation will destroy any data currently on the
disk.
The disk initialization failed, probably because the disk is damaged.
All instrument state registers have been cleared from memory along with any
saved calibration data, memory traces , and calibration kit de nitions .
A dditionally , all user-settable selections (such as HP-IB addresses) are set to
their defaults .
Y our last front panel or HP-IB request could not be implemented due to
insucient memory space . In some cases , this is a fatal error from which you
can escape only by presetting the instrument.
There is not enough memory available for the print or plot function. Increase
the available memory by changing or eliminating a memory-intensive operation
such as reducing the number of points in the sweep .

There is not enough memory space for the power meter calibration array .
Increase the available memory by clearing one or more save/recall registers , or
by reducing the number of points .
Y ou pressed an unde ned softkey .
List mode has been turned o in the frequency o set mode because it is
incompatible with your selected LO frequency .
The frequency list is empty . T o implement list frequency mode , add segments
to the list table .
A logarithmic sweep is only valid if the stop frequency is greater than four
times the start frequency . F or frequency spans of less than two octaves , the
sweep type automatically reverts to linear sweep .
The frequency domain data points must be harmonically related from dc to the
stop frequency . That is , stop = n start, where n = number of points . If this
condition is not true when a low pass mode (step or impulse) is selected and
transform is turned on, the analyzer resets the start and stop frequencies . The
stop frequency is set close to the entered stop frequency , and the start
frequency is set equal to stop/n.
All the memory in the sequence you are modifying is lled with instrument
commands .

When you use a sliding load (in a user-de ned calibration kit), you must set at
least three slide positions to complete the calibration.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RESUME CAL SEQUENCE softkey is not valid unless a calibration is already
in progress . Start a new calibration.
Y ou have no disk in the current disk unit. Insert a disk, or check the disk unit
number stored in the analyzer .
The self-diagnose function of the instrument operates on an internal test
failure . At this time , no failure has been detected.
No les of the type created by an analyzer store operation were found on the
disk or the disk drive is empty . If you requested a speci c le title , that le
was not found on the disk.
The rst IF signal was not detected during pretune . Check the front panel R
channel jumper . If there is no visible problem with the jumper , refer to the
HP 8753D Network Analyzer Service Guide for troubleshooting.
Y ou can turn limit lines on but they cannot be displayed on polar or Smith
chart display formats .

Y ou must turn the delta marker mode on, with at least two markers displayed,
in order to use the
MARKER -> SPAN softkey function.
Y ou cannot perform interpolated error correction due to insucient memory .
Y ou cannot modify the sequence due to insucient memory .
The rst IF signal was detected at pretune , but phase lock could not be
acquired. Check the signal level to the R channel input to make sure it is 35
dBm or higher . Refer to the HP 8753D Network Analyzer Service Guide for
troubleshooting.
Y ou cannot store a calibration set due to insucient memory . Y ou can free
more memory by clearing a saved instrument state from an internal register
(which may also delete an associated calibration set, if all the instrument states
using the calibration kit have been deleted). Y ou can store the saved
instrument state and calibration set to a disk before clearing them. After
deleting the instrument states , press
to run the memory packer .
When the analyzer is performing a power meter calibration, the HP-IB bus is
unavailable for other functions such as printing or plotting.
The store operation will over ow the available disk space . Insert a new disk or
purge les to create free disk space .

If you are going to display or otherwise use a memory trace , you must rst
store a data trace to memory .
Y ou have requested the analyzer , over HP-IB (or by sequencing), to load an
instrument state from an empty internal register .
Y ou can only use alpha-numeric characters (and underscores) in disk le titles
or internal save register titles . Other symbols are not allowed, except for the
\underscore" symbol.
The function you requested requires a capability provided by an option to the
standard analyzer . That option is not currently installed. (Refer to Chapter 1
for a description of the available options .)
See error number 57.
See error number 57.
Y ou have exceeded approximately +14 dBm at one of the test ports . The RF
output power is automatically reduced to 85 dBm. The annotation P+ appears
in the left margin of the display to indicate that the power trip function has
been activated. When this occurs , reset the power to a lower level, then toggle
SOURCE PWR on OFF softkey to switch on the power again.

Y ou have de ned the parallel port as COPY for sequencing in the HP-IB menu.
T o access the parallel port for general purpose I/O (GPIO), set the selection to
[GPIO].
Y ou have de ned the parallel port as general purpose I/O (GPIO) for
sequencing. The de nition was made under the
key menus . T o access
the parallel port for copy , set the selection to
PARALLEL [COPY] .
An internal phase lock calibration routine is automatically executed at
power-on, preset, and any time a loss of phase lock is detected. This message
indicates that phase lock calibration was initiated and the rst IF detected, but
a problem prevented the calibration from completing successfully . Refer to the
HP 8753D Network Analyzer Service Guide and execute pretune correction test
This message may appear if you connect a mixer between the RF output and R
input before turning on frequency o set mode . Ignore it: it will go away when
you turn on frequency o set. This message may also appear if you turn on
frequency o set mode before you de ne the o set.
Phase lock was acquired but then lost. Refer to the HP 8753D Network
Analyzer Service Guide for troubleshooting information.
When you press the
key , the analyzer aborts the plot in progress .
The plotter does not respond to control. V erify power to the plotter , and check
the HP-IB connection between the analyzer and the plotter . Ensure that the
plotter address recognized by the analyzer matches the HP-IB address set on
the plotter itself.

The plotter pinch wheels clamp the paper in place . If you raise the pinch
wheels , the plotter indicates a \not ready" status on the bus .
Phase lock has been achieved, but the source may be phase locked to the
wrong harmonic of the synthesizer . P erform the source pretune correction
routine documented in the \A djustments and Correction Constants" chapter in
the HP 8753D Network Analyzer Service Guide.
The power meter indicates an out-of-range condition. Check the test setup .
Sequential power meter readings are not consistent. V erify that the equipment
is set up correctly . If so , preset the instrument and restart the operation.
The temperature sensors on the A8 post-regulator assembly have detected an
over-temperature condition. The power supplies regulated on the
post-regulator have been shut down.
One or more supplies on the A8 post-regulator assembly have been shut down
due to an over-current, over-voltage , or under-voltage condition.
There is either a hardware failure in the source or you have attempted to set
the power level too high. The analyzer allows the output power to be set
higher or lower than the speci ed available power range . However , these
output powers may be unleveled or unavailable . Check to see if the power
level you set is within speci cations . If it is , refer to the HP 8753D Network
Analyzer Service Guide for troubleshooting.

PRESS [MENU], SELECT CW (IF) FREQ, THEN SWEPT LO
When you are sweeping the RF and LO , the IF must be xed.
When you press the
key , the analyzer aborts output to the printer .
print color not supported with EPSON
Y ou have de ned the printer type as EPSON-P2. Color print is not supported
with this printer . The print will abort.
The parallel port printer is not accepting data.
The parallel port printer is malfunctioning. The analyzer cannot complete the
copy function.
There is no printer connected to the parallel port.
The printer at the parallel port is not responding.
The printer at the parallel port is not set on line .

The printer does not respond to control. V erify power to the printer , and check
the HP-IB connection between the analyzer and the printer . Ensure that the
printer address recognized by the analyzer matches the HP-IB address set on
the printer itself.
There is a paper-related problem with the parallel port printer such as a paper
jam or out-of-paper condition.
The power to the printer at the parallel port is o .
If a print or plot is in progress and you attempt a second print or plot, this
message is displayed and the second attempt is ignored. T o abort a print or plot
in progress , press
.
One or both of the probe power supplies have been shut down due to an
over-current, over-voltage , or under-voltage condition.
The display information is being processed for a screen print to a copy device
and stored in the copy spool bu er . During this time , the analyzer's resources
are dedicated to this task (which takes less than a few seconds).

The power meter cannot be accessed by the analyzer . V erify that the power
meter address and model number set in the analyzer match the address and
model number of the actual power meter .
The analyzer does not currently contain the data you have requested. F or
example , this condition occurs when you request error term arrays and no
calibration is active .
Y ou cannot store an instrument state in an internal register due to insucient
memory . Increase the available memory by clearing one or more save/recall
registers and pressing
, or by storing les to a disk.
SELECTED SEQUENCE IS EMPTY
The sequence you attempted to run does not contain instrument commands .
SELF TEST #n FAILED
Internal test #n has failed. Several internal test routines are executed at
instrument preset. The analyzer reports the rst failure detected. Refer to the
HP 8753D Network Analyzer Service Guide for troubleshooting information on
internal tests and the self-diagnose feature .
The sequence running was stopped prematurely when you pressed the
key .
When you pause a sequence , you cannot continue it if you have modi ed it.
Y ou must start the sequence again.

Y ou cannot perform sliding load measurements due to insucient memory .
Increase the available memory by clearing one or more save/recall registers and
, or by storing les to a disk.
SOURCE POWER TURNED OFF, RESET UNDER POWER MENU
Y ou have exceeded the maximum power level at one of the inputs and power
has been automatically reduced. The annotation P+ indicates that power trip
has been activated. When this occurs , reset the power and then press
NNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
POWER
SOURCE PWR on OFF , to switch on the power .
The analyzer is beginning to output data from the spool bu er to the copy
device . The analyzer resumes normal operation; the data is being output to the
copy device in the background.
STOP/CW FREQ + OFFSET MUST BE < 3 GHz
The output frequency of the mixer cannot violate the minimum/maximum
frequency of the analyzer .
If you select external source auto or manual instrument mode and you do not
also select CW mode , the analyzer is automatically switched to CW .
Y ou have made instrument changes that cause the analyzer sweep time to be
automatically increased. Some parameter changes that cause an increase in
sweep time are narrower IF bandwidth, an increase in the number of points ,
and a change in sweep type .

The fractional-N and digital IF circuits have lost synchronization. Refer to the
HP 8753D Network Analyzer Service Guide for troubleshooting information.
The instrument is in a hold state and is no longer sweeping. T o take a new
NNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
sweep , press
TRIGGER MENU
CONTINUOUS .
Y ou have improperly formatted an HP-IB command. Refer to the HP 8753D
Network Analyzer Programmer's Guide for proper command syntax.
The analyzer is in talker/listener mode . In this mode , the analyzer cannot
control a peripheral device on the bus . Use the local menu to change to system
controller or pass control mode .
Y ou have prematurely stopped a service test.
Y ou have set frequencies in the list that are outside of the allowable frequency
range for harmonic measurements , or are greater than 3 GHz on instruments
without Option 006. Reduce the frequency range of the list.

Y ou can only nest sequences to a maximum level of six. The sequence will
abort if you nest more than six.
Y ou can have a maximum of 30 segments or 1632 points in frequency list mode .
In power meter calibrations , you can have a maximum of 12 segments for
power sensor cal factors and power loss functions .
Y ou can perform a time domain transformation only in linear and CW sweep
types .
Y our equipment setup for the adjustment procedure in progress is not correct.
Check the setup diagram and instructions HP 8753D Network Analyzer Service
Guide. Start the procedure again.
In single sweep mode , the instrument ensures that all changes to the
instrument state , if any , have been implemented before taking the sweep . The
command that you have initiated is being processed and will not be complete
until the new sweep is completed. An asterisk * is displayed in the left margin
until a complete fresh sweep has been taken.
This message is displayed between the start and nish of a read or write
operation to a disk.

Y ou have instructed the analyzer to use pass control (USEP ASC). When you
send the analyzer an instruction that requires active controller mode , the
analyzer requests control of the bus and simultaneously displays this message .
If the message remains , the system controller is not relinquishing the bus .
Y ou have sent the data header \#A" to the analyzer with no preceding input
command (such as INPUD A T A). The instrument recognized the header but did
not know what type of data to receive . Refer to the HP 8753D Network
Analyzer Programmer's Guide for command syntax information.
Y ou have attempted to store , load, or read le titles , but your disk format does
not conform to the Logical Interchange F ormat (LIF) or DOS format. Y ou must
initialize the disk before reading or writing to it.

Error Messages in Numerical Order
Refer to the alphabetical listing for explanations and suggestions for solving the problems .
2 INV ALID KEY
3 CORRECTION CONST ANTS NOT STORED
4 PHASE LOCK CAL F AILED
5 NO IF FOUND: CHECK R INPUT LEVEL
6 POSSIBLE F ALSE LOCK
7 NO PHASE LOCK: CHECK R INPUT LEVEL
8 PHASE LOCK LOST
9 LIST T ABLE EMPTY
10 CONTINUOUS SWITCHING NOT ALLOWED
11 SWEEP TIME INCREASED
12 SWEEP TIME TOO F AST
13 A VERA GING INV ALID ON NON-RA TIO MEASURE
15 NO MARKER DEL T A - SP AN NOT SET
17 DEMODULA TION NOT V ALID
list freq. table
22 POWER SUPPL Y SHUT DOWN!
23 PROBE POWER SHUT DOWN!

28 PLOTTER NOT READY-PINCH WHEELS UP
30 REQUESTED D A T A NOT CURRENTL Y A V AILABLE
31 ADDRESSED TO T ALK WITH NOTHING TO SA Y
33 SYNT AX ERROR
34 BLOCK INPUT ERROR
35 BLOCK INPUT LENGTH ERROR
36 SYST CTRL OR P ASS CTRL IN LOCAL MENU
40 DISK MEDIUM NOT INITIALIZED
41 NO DISK MEDIUM IN DRIVE
42 FIRST CHARA CTER MUST BE A LETTER
43 ONL Y LETTERS AND NUMBERS ARE ALLOWED
44 NOT ENOUGH SP A CE ON DISK FOR STORE
46 ILLEGAL UNIT OR V OLUME NUMBER
47 INITIALIZA TION F AILED
48 DISK IS WRITE PROTECTED
49 DISK WEAR-REPLA CE DISK SOON
50 TOO MANY SEGMENTS OR POINTS
51 INSUFFICIENT MEMORY
54 NO V ALID MEMORY TRA CE
55 NO V ALID ST A TE IN REGISTER

57 O VERLO AD ON INPUT R, POWER REDUCED
58 O VERLO AD ON INPUT A, POWER REDUCED
59 O VERLO AD ON INPUT B , POWER REDUCED
63 CALIBRA TION REQUIRED
64 CURRENT P ARAMETER NOT IN CAL SET
65 CORRECTION AND DOMAIN RESET
66 CORRECTION TURNED OFF
67 DOMAIN RESET
68 ADDITIONAL ST AND ARDS NEEDED
69 NO CALIBRA TION CURRENTL Y IN PROGRESS
70 NO SP A CE FOR NEW CAL. CLEAR REGISTERS
72 EXCEEDED 7 ST AND ARDS PER CLASS
73 SLIDES ABORTED (MEMORY REALLOCA TION)
74 CALIBRA TION ABORTED
75 FORMA T NOT V ALID FOR MEASUREMENT
77 WRONG DISK FORMA T , INITIALIZE DISK
111 DEADLOCK
112 SELF TEST #n F AILED
114 NO F AIL FOUND
115 TROUBLE! CHECK SETUP AND ST ART O VER
117 PWR MTR: NOT ON/CONNECTED OR WRONG ADDRS
118 POWER METER NOT SETTLED
123 NO MEMORY A V AILABLE FOR INTERPOLA TION

126 NO MEMORY A V AILABLE FOR SEQUENCING
127 CAN'T STORE/LO AD SEQUENCE, INSUFFICIENT MEMORY
130 D2/D1 INV ALID WITH SINGLE CHANNEL
131 FUNCTION NOT V ALID DURING MOD SEQUENCE
132 MEMORY FOR CURRENT SEQUENCE IS FULL
133 THIS LIST FREQ INV ALID IN HARM/3 GHZ RNG
140 FREQ OFFSET ONL Y V ALID IN NETWORK ANAL YZER MODE
141 STOP/CW FREQ + OFFSET MUST BE < 3 GHz
148 EXTERNAL SOURCE MODE REQUIRES CW TIME
150 LOG SWEEP REQUIRES 2 OCT A VE MINIMUM SP AN
153 SEQUENCE MA Y HA VE CHANGED , CAN'T CONTINUE
157 SEQUENCE ABORTED
159 CH1 (CH2) T ARGET V ALUE NOT FOUND
161 PRESS [MENU], SELECT CW (IF) FREQ, THEN SWEPT LO
163 FUNCTION ONL Y V ALID DURING MOD SEQUENCE
165 P ARALLEL PORT NOT A V AILABLE FOR GPIO
167 P ARALLEL PORT NOT A V AILABLE FOR COPY
169 HPIB COPY IN PROGRESS, ABORT WITH LOCAL
170 COPY :device not responding; copy aborted

178 print color not supported with EPSON
180 DOS NAME LIMITED TO 8 CHARS + 3 CHAR EXTENSION
181 B AD FREQ FOR HARMONIC OR FREQ OFFSET
183 B A TTERY F AILED . ST A TE MEMORY CLEARED
184 B A TTERY LOW! STORE SA VE REGS TO DISK
185 CANNOT FORMA T DOS DISKS ON THIS DRIVE
188 DIRECTORY FULL
189 DISK READ/WRITE ERROR
190 DISK MESSA GE LENGTH ERROR
191 EXT SOURCE NOT READY FOR TRIGGER
192 FILE NOT FOUND

198 NOT ALLOWED DURING POWER METER CAL
199 CANNOT MODIFY F A CTORY PRESET
200 ALL REGISTERS HA VE BEEN USED
201 FUNCTION NOT V ALID FOR INTERNAL MEMORY
202 FUNCTION NOT A V AILABLE
203 CANNOT READ/WRITE HFS FILE SYSTEM
204 FREQS CANNOT BE CHANGED , TOO MANY POINTS



This chapter contains the following information:
Connecting and con guring peripherals
Where to Look for More Information
A dditional information about many of the topics discussed in this chapter is located in the
following areas:
Chapter 2, \Making Measurements ," contains step-by-step procedures for making
measurements or using particular functions .
Chapter 6, \Application and Operation Concepts ," contains explanatory-style information
about many applications and analyzer operation.
Measurement A ccessories A vailable
The following calibration kits contain precision standards and required adapters of the
indicated connector type . The standards (known devices) facilitate measurement calibration,
also called vector error correction. Refer to the data sheet and ordering guide for additional
information. P arts numbers for the standards are in their manuals .
HP 85031B 7 mm Calibration Kit
HP 85032B 50 Ohm Type-N Calibration Kit
HP 85033D 3.5 mm Calibration Kit
HP 85033C 3.5 mm Calibration Kit
HP 85036B 75 Ohm Type-N Calibration Kit
HP 85039A 75 Ohm Type-F Calibration Kit
The HP 85033D is the recommended 3.5 mm calibration kit as it provides
greater measurement accuracy than the HP 85033C and it is easier to use due
to one-piece opens .
Compatible P eripherals 11-1

V eri cation Kit
A ccurate operation of the analyzer system can be veri ed by measuring known devices other
than the standards used in calibration, and comparing the results with recorded data.
HP 85029B 7 mm V eri cation Kit
This kit contains traceable precision 7 mm devices used to con rm the system's error-corrected
measurement uncertainty performance . Also included is veri cation data on a 3.5 inch disk,
together with a hard-copy listing. A system veri cation procedure is provided with this kit and
also in the HP 8753D Service Guide.
T est P ort Return Cables
The following RF cables are used to connect a two-port device between the test ports . These
cables provide shielding for high dynamic range measurements .
HP 11857D 7 mm T est P ort Return Cable Set
This set consists of a pair of test port return cables that can be used in measurements of 7 mm
devices . They can also be used with connectors other than 7 mm by using the appropriate
precision adapters .
HP 11857B 75 Ohm Type-N T est P ort Return Cable Set
This set consists of test port return cables for use with the HP 8753D Option 075.
HP 11852B 50 to 75 Ohm Minimum Loss P ad.
This device converts impedance from 50 ohms to 75 ohms or from 75 ohms to 50 ohms . It is
used to provide a low SWR impedance match between a 75 ohm device under test and the
HP 8753D network analyzer (without Option 075).
These adapter kits contain the connection hardware required for making measurements on
devices of the indicated connector type .
HP 11853A 50 Ohm Type-N A dapter Kit
HP 11854A 50 Ohm BNC A dapter Kit
HP 11855A 75 Ohm Type-N A dapter Kit
HP 11856A 75 Ohm BNC A dapter Kit
11-2 Compatible P eripherals
A dapter Kits

Transistor T est Fixtures
The following Hewlett-P ackard transistor test xtures are compatible with the HP 8753D .
A dditional test xtures for transistors and other devices are available from Inter-Continental
Microwave . T o order their catalog, request HP literature number 5091{4254E. Or contact
Inter{Continental Microwave as follows:
Santa Clara, CA 95054-1524
HP 11600B and 11602B Transistor Fixtures .
These xtures are used to hold devices for S-parameter measurements in a 50 ohm coaxial
circuit. They can be used to measure bipolar or eld-e ect transistors in several con gurations ,
from dc to 2.0 GHz. The HP 11600B accepts transistors with TO-18 to TO-72 package
dimensions , and the HP 11602B accepts transistors with TO-5 to TO-12 package dimensions .
Both xtures can also be used to measure other circuit elements such as diodes , resistors , or
inductors , which have 0.016 to 0.019 inch diameter leads .
HP 11608A Option 003 Transistor Fixture .
This xture is designed to be user-milled to hold stripline transistors for S-parameter
measurements . Option 003 is pre-milled for 0.205 inch diameter disc packages , such as the
HP HP A C-200.
HP 11858A Transistor Fixture A dapter .
This adapter provides a rigid RF cable interconnection between the HP 8753D test port
connectors and the HP 11600B , 11602B , or 11608A transistor xture .
Compatible P eripherals 11-3
1515 Wyatt Drive · (tel) 408 727-1596 · (fax) 408 727-0105

System A ccessories A vailable
The HP 85043D system cabinet is designed to rack mount the analyzer in a system
con guration. The 132 cm (52 in) system cabinet includes a bookcase , a drawer , and a
convenient work surface .
The HP 1181A system testmobile is designed to provide mobility for instruments , test systems ,
and work stations . This system testmobile can hold units up to 610-mm (24 in) deep and has a
load capacity of up to 90 kg (200 lb) on the tilt tray and 227 kg (500 lb) total.
The analyzer is capable of plotting or printing displayed measurement results directly (without
the use of an external computer) to a compatible peripheral. The analyzer supports HP-IB ,
serial, and parallel peripherals . Most Hewlett-P ackard desktop printers and plotters are
compatible with the analyzer . Some common compatible peripherals are listed here (some are
no longer available for purchase but are listed here for your reference):
These plotters are compatible:
HP 7440A ColorPro Eight-P en Color Graphics Plotter
HP 7470A Two-P en Graphics Plotter
HP 7475A Six-P en Graphics Plotter
HP 7550A/B High-Speed Eight-P en Graphics Plotter
These printers are compatible:
HP C2621A, DeskJet P ortable 310
HP C2634A, DeskJet P ortable 320
HP C2655A, DeskJet P ortable 340
HP C2642A, DeskJet 400
HP C2106A, DeskJet 500
HP C2114A, DeskJet 500C
HP C2170A, DeskJet 520
HP C2162A, DeskJet 540
HP C2121A, DeskJet 550C
HP C2168A, DeskJet 560C
HP C2184A, DeskJet 600
HP C2164A, DeskJet 660C
HP C4567A, DeskJet 682C
HP C4562A, DeskJet 690C
HP C2145A, DeskJet 850C
11-4 Compatible P eripherals
System Cabinet · System T estmobile · Plotters and Printers

HP C1676A, DeskJet 1200C (can also be used to plot)
HP C3540A, DeskJet 1600C (can also be used to plot)
All LaserJets (can also be used to plot)
HP 3630A P aintJet Color Graphics Printer
Epson printers which are compatible with the Epson ESC/P2 printer control language , such
as the LQ570, are also supported by the analyzer . Older Epson printers , however , such as the
FX-80, will not work with the analyzer .
The analyzer has the capability of storing instrument states directly to its internal memory ,
to an internal disk drive , or to an external disk drive . The internal 3.5 inch oppy disk can
be initialized in both LIF and DOS formats and is capable of reading and writing data in both
formats . Using the internal disk drive is the preferred method, but the capability to use
external disk drives still exists . Most external disks using CS80 protocol are compatible .
The analyzer does not support the LIF-1 (hierarchy le system) directory
format.
Do not use the older single-sided disks in the analyzer's internal drive .
An HP-IB cable is required for interfacing the analyzer with a plotter , printer , external disk
drive , or computer . The cables available are:
HP 10833A HP-IB Cable , 1.0 m (3.3 ft.)
HP 10833B HP-IB Cable , 2.0 m (6.6 ft.)
HP 10833D HP-IB Cable , 0.5 m (1.6 ft.)
HP C2912B Centronics (P arallel) Interface Cable , 3.0 m (9.9 ft.)
HP C2913A RS-232C Interface Cable , 1.2 m (3.9 ft.)
HP C2914A Serial Interface Cable , 1.2 m (3.9 ft.)
HP 24542G Serial Interface Cable , 3 m (9.9 ft.)
HP 24542D P arallel Interface Cable , 2 m (6 ft.)
HP 92284A P arallel Interface Cable , 2 m (6 ft.)
Compatible P eripherals 11-5

A keyboard can be connected to the analyzer for data input such as titling les .
A keyboard can be connected to the analyzer for data input such as titling les . The
HP C1405A Option AB A keyboard with the HP part number C1405-60015 adapter is suitable for
this purpose . Or , the analyzer is designed to accept most PC-A T-compatible keyboards with a
standard DIN connector . Keyboards with a mini-DIN connector are compatible with the HP part
number C1405-60015 adapter .
An external controller is not required for measurement calibration or time domain capability .
However , some performance test procedures are semi-automated and require the use of an
external controller . (The system veri cation procedure does not require an external controller .)
The system can be automated with the following computers:
IBM PC compatible computer with an HP-IB/GP-IB interface card
HP 9000 series 200/300 workstation
HP 9000 series 700 workstation with HP B ASIC-UX
F or more information about compatible computers , call your nearest Hewlett-P ackard Sales and
Service oce .
A set of sample measurement programs is provided with the HP 8753D Network Analyzer
Programmer's Guide on a 3.5 inch disk. The programs include typical measurements to be
used as an introductory example for programming the analyzer over HP-IB . It is designed to be
easily modi ed for use in developing programs for speci c needs . The programs are compatible
with HP B ASIC versions 2.0 and later , QuickBasic , and QuickC, and will run on an IBM PC
compatible computer (as well as the HP 9000 series 200/300/700 workstations), using any HP
8753D compatible printer or plotter .
The following additional software products are also available for the HP 8753D:
HP 85190A IC-CAP Modelling Suite
(MDS) HP Microwave Design System
The analyzer can drive both its internal display and an external monitor simultaneously . One
recommended color monitor is the HP 35741A/B .
External Monitor Requirements:
60 Hz vertical refresh rate
25.5 kHz horizontal refresh rate
RGB with synchronization on green
75 ohm video input impedance
video amplitude 1 Vp-p (0.7 V = white , 0 V = black, 0.3 V = synchronization)
11-6 Compatible P eripherals

Connecting the P eripheral Device
Connect the peripheral to the corresponding interface port.
Printer Interface Recommended Cable
P arallel HP 92284A
Figure 11-1. P eripheral Connections to the Analyzer
The keyboard can be connected to the analyzer while the power is on or o .
Compatible P eripherals 11-7
Connecting P eripherals · Serial HP 24542G

Con guring the Analyzer for the P eripheral
All copy con guration settings are stored in non-volatile memory . Therefore , they are not
a ected if you press
or switch o the analyzer power .
If the P eripheral is a Printer
SET ADDRESSES
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PRINTER PORT
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PRNTR TYPE until the correct printer choice
appears:
NNNNNNNNNNNNNNNNNNNNNNN
DeskJet (except for HP DeskJet 540 and DeskJet 850C)
NNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNN
Epson-P2 (printers that conform to the ESC/P2 printer control language)
DJ 540 (for use with the HP DeskJet 540 and DeskJet 850C)
DJ 540 converts 100 dpi raster information to 300 dpi raster format.
If your DeskJet printer does not support the 100 dpi raster format and your
printing results seem to be less than normal size , select
DJ 540 .
PRNTR PORT HPIB if your printer has an HP-IB interface , and then con gure the
print function as follows:
a. Enter the HP-IB address of the printer , followed by
.
b.
LOCAL
SYSTEM CONTROLLER if there is no external controller connected to
the HP-IB bus .
c. Press
USE PASS CONTROL if there is an external controller connected to
the HP-IB bus .
PARALLEL if your printer has a parallel (centronics) interface , and then con gure
the print function as follows:
and then select the parallel port interface function by pressing
PARALLEL
LOCAL
until the correct function appears:
PARALLEL [COPY] , the parallel port is dedicated for normal copy
device use (printers or plotters).
PARALLEL [GPIO] , the parallel port is dedicated for general purpose
I/O , and cannot be used for printing or plotting.
11-8 Compatible P eripherals

SERIAL if your printer has a serial (RS-232) interface , and then con gure the print
function as follows:
PRINTER BAUD RATE and enter the printer's baud rate , followed by
.
a. Press
b. T o select the transmission control method that is compatible with your printer , press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
XMIT CNTRL (transmit control - handshaking protocol) until the correct method
appears:
Xon-Xoff , the handshake method allows the printer to control the
data exchange by transmitting control characters to the network analyzer .
DTR-DSR , the handshake method allows the printer to control the data
exchange by setting the electrical voltage on one line of the RS-232 serial cable .
DTR-DSR handshake takes place in the hardware rather than the
rmware or software , it is the fastest transmission control method.
If the P eripheral is a Plotter
HPGL/2 Compatible Printer (used as a plotter)
SET ADDRESSES
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PRINTER PORT and then press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PRNTR TYPE until the correct
printer choice appears:
NNNNNNNNNNNNNNNNNNNNNNN
DeskJet (only DeskJet 1200C and DeskJet 1600C)
LaserJet (only LaserJet III and IV)
NNNNNNNNNNNNNNNNNNNNNNNNNN
Epson-P2 (printers that conform to the ESC/P2 printer control language)
PRNTR PORT HPIB if your printer has an HP-IB interface , and then con gure the
print function as follows:
a. Enter the HP-IB address of the printer (default is 01), followed by
.
b.
LOCAL
SYSTEM CONTROLLER if there is no external controller connected to
the HP-IB bus .
c. Press
USE PASS CONTROL if there is an external controller connected to
the HP-IB bus .
Compatible P eripherals 11-9

PARALLEL if your printer has a parallel (centronics) interface , and then con gure
the print function as follows:
LOCAL
and then select the parallel port interface function by pressing
until the correct function appears:
PARALLEL [COPY] , the parallel port is dedicated for normal copy
device use (printers or plotters).
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PARALLEL [GPIO] , the parallel port is dedicated for general purpose
I/O , and cannot be used for printing or plotting.
NNNNNNNNNNNNNNNNNNN
SERIAL if your printer has a serial (RS-232) interface , and then con gure the print
function as follows:
a. Press
PRINTER BAUD RATE and enter the printer's baud rate , followed by
.
b. T o select the transmission control method that is compatible with your printer , press
XMIT CNTRL (transmit control - handshaking protocol) until the correct method
appears:
NNNNNNNNNNNNNNNNNNNNNNNNN
Xon-Xoff , the handshake method allows the printer to control the
data exchange by transmitting control characters to the network analyzer .
DTR-DSR , the handshake method allows the printer to control the data
exchange by setting the electrical voltage on one line of the RS-232 serial cable .
NNNNNNNNNNNNNNNNNNNNNN
DTR-DSR handshake takes place in the hardware rather than the
rmware or software , it is the fastest transmission control method.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SET ADDRESSES
PLOTTER PORT and then
PLTR TYPE [HPGL PRT] appears .
11-10 Compatible P eripherals

NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SET ADDRESSES
PLOTTER PORT and then
PLTR TYPE [PLOTTER] appears .
PLTR PORT HPIB if your plotter has an HP-IB interface , and then con gure the
plot function as follows:
a. Enter the HP-IB address of the printer (default is 05), followed by
.
b.
LOCAL
SYSTEM CONTROLLER if there is no external controller connected to
the HP-IB bus .
c. Press
USE PASS CONTROL if there is an external controller connected to
the HP-IB bus .
PARALLEL if your printer has a parallel (centronics) interface , and then con gure
the print function as follows:
and then select the parallel port interface function by pressing
PARALLEL
LOCAL
until the correct function appears:
PARALLEL [COPY] , the parallel port is dedicated for normal copy
device use (printers or plotters).
PARALLEL [GPIO] , the parallel port is dedicated for general purpose
I/O , and cannot be used for printing or plotting.
SERIAL if your printer has a serial (RS-232) interface , and then con gure the print
function as follows:
PRINTER BAUD RATE and enter the printer's baud rate , followed by
.
a. Press
b. T o select the transmission control method that is compatible with your printer , press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
XMIT CNTRL (transmit control - handshaking protocol) until the correct method
appears:
Xon-Xoff , the handshake method allows the printer to control the
data exchange by transmitting control characters to the network analyzer .
DTR-DSR , the handshake method allows the printer to control the data
exchange by setting the electrical voltage on one line of the RS-232 serial cable .
DTR-DSR handshake takes place in the hardware rather than the
rmware or software , it is the fastest transmission control method.
Compatible P eripherals 11-11

If the P eripheral is a P ower Meter
SET ADDRESSES .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
HP 437B or 438A
a. Enter the HP-IB address of the power meter , followed by
.
b.
LOCAL
SYSTEM CONTROLLER if there is no external controller connected to the
HP-IB bus .
c. Press
USE PASS CONTROL if there is an external controller connected to the
HP-IB bus .
If the P eripheral is an External Disk Drive
DISK UNIT NUMBER and enter the drive where your disk is located, followed
.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
If your storage disk is partitioned, press
VOLUME NUMBER and enter the volume number
where you want to store the instrument state le .
SET ADDRESSES
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ADDRESS: DISK .
and select one of the following:
SYSTEM CONTROLLER if there is no external controller connected.
PASS CONTROL when an external controller is connected and the analyzer needs
to control peripherals directly . This mode causes the analyzer to request control from the
external controller whenever it needs to communicate with a peripheral.
If the P eripheral is a Computer Controller
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LOCAL
SET ADDRESSES
ADDRESS: CONTROLLER .
and select one of the following:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TALKER/LISTENER to allow the computer controller to be involved in all
peripheral access operations .
PASS CONTROL when an external controller is connected and the analyzer needs
to control peripherals directly . This mode causes the analyzer to request control from the
external controller whenever it needs to communicate with a peripheral.
11-12 Compatible P eripherals

Con guring the Analyzer to Produce a Time Stamp
Y ou can set a clock, and then activate it, if you want the time and date to appear on your
hardcopies .
SYSTEM
SET CLOCK .
SET YEAR and enter the current year , followed by
.
SET MONTH and enter the current month of the year , followed
.
SET DAY and enter the current day of the month, followed by
.
SET HOUR and enter the current hour of the day (0-23), followed by
.
SET MINUTES and enter the next immediate minute , followed by
.
ROUND SECONDS when the current time is exactly as you have set it.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TIME STAMP ON appears on the softkey label.
Compatible P eripherals 11-13

The analyzer is factory-equipped with a remote programming digital interface using the
Hewlett-P ackard Interface Bus (HP-IB). HP-IB is Hewlett-P ackard's hardware , software ,
documentation, and support for IEEE 488.1 and IEC-625, worldwide standards for interfacing
instruments . The HP-IB lets you control the analyzer with an external computer that sends
commands or instructions to and receives data from the analyzer . This provides a remote
operator with the same control of the instrument available to the local operator , except for
control of the power line switch and some internal tests .
In addition, without the use of an external computer , the analyzer can use HP-IB to output
measurement results directly to a compatible printer or plotter and to store data to an external
disk drive . It can also control a power meter for power calibration and, through a subset of
HP-GL (Hewlett-P ackard Graphics Language), user graphics can be plotted on the analyzer
display .
F or more complete information on programming the analyzer remotely over HP-IB , refer to
the HP 8753D Network Analyzer Programmer's Guide.
F or a complete general description of the HP-IB , refer to the Tutorial Description of the
Hewlett-P ackard Interface Bus, HP publication 5952-0156 and to Condensed Description of
the Hewlett-P ackard Interface Bus (HP part number 59401-90030).
F or more information on the IEEE-488.1 standard, refer to IEEE Standard Digital Interface
for Programmable Instrumentation, published by the Institute of Electrical and Electronics
Engineers , Inc ., 345 East 47th Street, New Y ork, New Y ork 10017.
11-14 Compatible P eripherals
HP-IB Programming Overview

The Hewlett-P ackard Interface Bus (HP-IB) is Hewlett-P ackard's hardware , software ,
documentation, and support for IEEE 488.2 and IEC-625 worldwide standards for interfacing
instruments . This interface allows you to operate the analyzer and peripherals in two methods:
by an external system controller
by the network analyzer in system-controller mode
The HP-IB employs a party-line bus structure in which up to 15 devices can be connected
on one contiguous bus . The interface consists of 16 signal lines and 8 ground lines within a
shielded cable . With this cabling system, many di erent types of devices including instruments ,
computers , power meters , plotters , printers , and disk drives can be connected in parallel.
Every HP-IB device must be capable of performing one or more of the following interface
functions:
A talker is a device capable of transmitting device-dependent data when addressed to talk.
There can be only one active talker at any given time . Examples of this type of device include:
The network analyzer is a talker when it sends trace data or marker information over the bus .
A listener is a device capable of receiving device-dependent data over the interface when
addressed to listen. There can be as many as 14 listeners connected to the interface at any
given time . Examples of this type of device include:
The network analyzer is a listener when it is controlled over the bus by a system controller .
A controller is de ned as a device capable of:
There can be only one active controller on the interface at any time . Examples of controllers
include desktop computers , minicomputers , workstations , and the network analyzer . In a
multiple-controller system, active control can be passed between controllers , but there can
only be one system controller connected to the interface . The system controller acts as the
master and can regain active control at any time . The analyzer is an active controller when it
plots , prints , or stores to an external disk drive in the pass-control mode . The analyzer is also a
system controller when it is operating in the system-controller mode .
Compatible P eripherals 11-15


Figure 11-2. HP-IB Bus Structure
The data bus consists of 8 bi-directional lines that are used to transfer data from one device to
another . Programming commands and data transmitted on these lines are typically encoded
in ASCII, although binary encoding is often used to speed up the transfer of large arrays .
Both ASCII- and binary-data formats are available to the analyzer . In addition, every byte
transferred over HP-IB undergoes a handshake to insure valid data.
A three-line handshake scheme coordinates the transfer of data between talkers and listeners .
T o insure data integrity in multiple-listener transfers , this technique forces data transfers
to occur at the transfer rate of the slowest device connected to the interface . With most
computing controllers and instruments , the handshake is performed automatically , making it
transparent to the programmer .
The data bus also has ve control lines . The controller uses these lines to address devices and
to send bus commands .
IFC (Interface Clear) This line is used exclusively by the system controller . When
this line is true (low), all devices (whether addressed or not)
unaddress and revert to an idle state .
11-16 Compatible P eripherals

A TN (Attention) The active controller uses this line to de ne whether
the information on the data bus is command-oriented or
data-oriented. When this line is true (low), the bus is in the
command mode , and the data lines carry bus commands . When
this line is false (high), the bus is in the data mode , and the
data lines carry device-dependent instructions or data.
SRQ (Service Request) This line is set true (low) when a device requests service
and the active controller services the requesting device . The
network analyzer can be enabled to pull the SRQ line for a
variety of reasons such as requesting control of the interface ,
for the purposes of printing, plotting, or accessing a disk.
REN (Remote Enable) This line is used exclusively by the system controller . When this
line is set true (low), the bus is in the remote mode , and devices
are addressed by the controller to either listen or talk. When
the bus is in remote mode and a device is addressed, it receives
instructions from the system controller via HP-IB rather than
from its front panel (pressing
returns the device to
front-panel operation). When this line is set false (high), the
bus and all of the connected devices return to local operation.
EOI (End or Identify) This line is used by a talker to indicate the last data byte in a
multiple-byte transmission, or by an active controller to initiate
a parallel-poll sequence . The analyzer recognizes the EOI line
as a terminator , and it pulls the EOI line with the last byte of a
message output (data, markers , plots , prints , error messages).
The analyzer does not respond to parallel poll.
Number of Interconnected Devices: 15 maximum.
Interconnection P ath Maximum Cable Length: 20 meters maximum or 2 meters per device
(whichever is less).
transfer using a 3-line handshake system.
speci ed distances with tri-state drivers . A ctual
data rate depends on the transfer rate of the
slowest device connected to the bus .
maximum of 1 talker and 14 listeners can be
connected to the interface at given time .
this instrument), only one controller can be
active at any given time . The active controller
can pass control to another controller , but only
the system controller can assume unconditional
control. Only one system controller is allowed.
Compatible P eripherals 11-17
HP-IB Requirements

On the network analyzer's rear panel, next to the HP-IB connector , there is a list of HP-IB
device subsets as de ned by the IEEE 488.2 standard. The analyzer has the following
capabilities:
SH1 Full-source handshake .
AH1 Full-acceptor handshake .
T6 Basic talker , answers serial poll, unaddresses if MLA is issued. No talk-only mode .
L4 Basic listener , unaddresses if MT A is issued. No listen-only mode .
SR1 Complete service request (SRQ) capabilities .
RL1 Complete remote/local capability including local lockout.
PP0 Does not respond to parallel poll.
DC1 Complete device clear .
DT1 Responds to a Group Execute Trigger (GET) in the hold-trigger mode .
C1,C2,C3 System controller capabilities in system-controller mode .
C10 P ass control capabilities in pass-control mode .
E2 Tri-state drivers .
LE0 No extended listener capabilities .
TE0 No extended talker capabilities .
These codes are completely explained in the IEEE Std 488 documents , published by the
Institute of Electrical and Electronic Engineers , Inc ., 345 East 47th Street, New Y ork,
New Y ork 11017.
11-18 Compatible P eripherals
HP-IB Operational Capabilities

When the analyzer is connected to other instruments over the HP-IB , the HP-IB status
indicators illuminate to display the current status of the analyzer . The HP-IB status indicators
are located in the instrument-state function block on the front panel of the network analyzer .
R = Remote Operation
L = Listen mode
T = T alk mode
S = Service request (SRQ) asserted by the analyzer
The analyzer uses a single-bus architecture . The single bus allows both the analyzer and the
host controller to have complete access to the peripherals in the system.
Three di erent controller modes are possible in and HP-IB system:

Figure 11-3. Analyzer Single Bus Concept
Compatible P eripherals 11-19

This mode allows the analyzer to control peripherals directly in a stand-alone environment
(without an external controller). This mode can only be selected manually from the
analyzer's front panel. It can only be used if no active computer or instrument controller is
connected to the system via HP-IB . If an attempt is made to set the network analyzer to the
system-controller mode when another controller is connected to the interface , the following
message is displayed on the analyzer's display screen:
The analyzer must be set to the system-controller mode in order to access peripherals from the
front panel. In this mode , the analyzer can directly control peripherals (plotters , printers , disk
drives , power meters , etc .) and the analyzer may plot, print, store on disk or perform power
meter functions .
Do not attempt to use this mode for programming. HP recommends using an
external instrument controller when programming. See the following section,
\T alker/Listener Mode ."
This is the mode that is normally used for remote programming of the analyzer . In
talker/listener mode , the analyzer and all peripheral devices are controlled from an external
instrument controller . The controller can command the analyzer to talk and other devices
to listen. The analyzer and peripheral devices cannot talk directly to each other unless the
computer sets up a data path between them. This mode allows the analyzer to act as either
a talker or a listener , as required by the controlling computer for the particular operation in
progress .
This mode allows the computer to control the analyzer via HP-IB (as with the talker/listener
mode), but also allows the analyzer to take control of the interface in order to plot, print, or
access a disk. During an analyzer-controlled peripheral operation, the host computer is free to
perform other internal tasks (i.e . data or display manipulation) while the analyzer is controlling
the bus . After the analyzer-controlled task is completed, the analyzer returns control to the
system controller .
Setting HP-IB A ddresses
In systems interfaced using HP-IB , each instrument on the bus is identi ed by an HP-IB
address . This address code must be di erent for each instrument on the bus . These addresses
are stored in short-term, non-volatile memory and are not a ected when you press
cycle the power . The analyzer occupies two HP-IB addresses: the instrument itself and the
display . The display address is derived from the instrument address by complementing the
instrument's least-signi cant bit. Hence , if the instrument is at an even address , the display
occupies the next higher address . If the instrument is at an odd address , the display occupies
the next lower address .
The analyzer addresses are set by pressing
SET ADDRESSES . In system-controller mode ,
the addresses must be set for the plotter , printer , disk drive , and power meter .
The default address for the analyzer is device 16, and the display address is device 17.
11-20 Compatible P eripherals

There is also an address for the system controller . This address refers to the
controller when the network analyzer is being used in pass-control mode . This
is the address that control is passed back to when the analyzer-controlled
operation is complete .
The analyzer HP-IB commands are derived from their front-panel key titles (where possible),
according to this naming convention:
Simple commands are the rst four letters of the function they control, as in POWE, the
command name for power . If the function label contains two words , the rst three mnemonic
letters are the rst three letters of the rst word, and the fourth mnemonic letter is the rst
letter of the second word. F or example , ELED is derived from electrical delay .
If there are many commands grouped together in a category , as in markers or plotting pen
numbers , the command is increased to 8 letters . The rst 4 letters are the category label
derived using rule 1. The last 4 letters are the function speci er , again derived using rule 1. As
an example , category pen numbers are represented by the command PENN, which is used in
combination with several functions such as PENND A T A, PENNMEMO .
The code naming guidelines , listed in T able 11-1, are used in order to:
make commands more meaningful and easier to remember
maintain compatibility with other products (including the HP 8510)
There are times when these guidelines are not followed due to technical
considerations .
Compatible P eripherals 11-21
Analyzer Command Syntax · Code Naming Convention

T able 11-1. Code Naming Convention
Convention K ey Title F or HP-IB Code Use Example
One W ord P ower First F our Letters POWE
Two W ords Electrical Delay First Three Letters of First W ord, First Letter
of Second W ord
Two W ords in a Group Marker !Center F our Letters of Both MARKCENT
Gate !Span GA TESP AN
Three W ords Cal Kit N 50 First Three Letters of First W ord, First Letter
of Second W ord, First F our Letters of Third
P en Num Data PENND A T A
Some codes require appendages (ON, OFF , 1, 2, etc .). Codes that do not have a front-panel
equivalent are HP-IB only commands . They use a similar convention based on the common
name of the function.
The analyzer accepts the following ASCII characters:
Both upper- and lower-case letters are acceptable . Carriage returns , leading zeros , spaces , and
unnecessary terminators are ignored, except for those within a command or appendage . If the
analyzer does not recognize a character as appropriate , it generates a syntax error message and
recovers at the next terminator .
11-22 Compatible P eripherals

The analyzer can input and output data in basic units such as Hz, dB , seconds , ohms , etc .
S Seconds HZ Hertz
V V olts DB dB or dBm
Input data is assumed to be in basic units unless one of the following units expressions quali es
the data input (upper and lower case are equivalent):
MS Milliseconds kHz Kilohertz
US Microseconds MHz Megahertz
NS Nanoseconds GHz Gigahertz
PS Picoseconds FS F emtoseconds
An HP-IB diagnostic feature (debug mode) is available in the HP-IB menu. A ctivating the debug
mode causes the analyzer to scroll incoming HP-IB commands across the display . Nonprintable
characters are represented with a . Any time the analyzer receives a syntax error , the
commands halt, and a pointer indicates the misunderstood character . See the HP 8753D
Network Analyzer Programmer's Guide for information on correct programming syntax.
Compatible P eripherals 11-23
Units · HP-IB Debug Mode



Preset State and Memory Allocation
The analyzer is capable of saving complete instrument states for later retrieval. It can store
these instrument states into the internal memory , to the internal disk, or to an external disk.
This chapter describes these capabilities in the following sections:
instrument state de nition
internal and external data storage
description of analyzer state after preset
Where to Look for More Information
A dditional information about many of the topics discussed in this chapter is located in the
following areas:
Chapter 2, \Making Measurements ," contains step-by-step procedures for making
measurements or using particular functions .
Chapter 4, \Printing, Plotting, and Saving Measurement Results ," contains instructions
for saving to disk or the analyzer internal memory , and printing and plotting displayed
measurements .
Types of Memory and Data Storage
The analyzer utilizes two types of memory and can also utilize the internal disk drive or be
connected to an external disk drive:
This is dynamic read/write memory , of approximately 2 Mbytes , that contains all of the
parameters that make up the current instrument state . An instrument state consists of all the
stimulus and response parameters that set up the analyzer to make a speci c measurement.
Some data that you may think is part of the instrument state (such as calibration data and
memory traces) are actually stored in non-volatile memory . See \Non-V olatile Memory" to read
more about the di erences .
V olatile memory is cleared upon a power cycle of the instrument and, except as noted, upon
instrument preset.
Preset State and Memory Allocation 12-1
memory allocation · V olatile Memory

This is CMOS read/write memory that is protected by a battery to provide storage of data when
line power to the instrument is turned o . With this battery protection, data can be retained in
memory for 250 days at 70
C and for 10 years at 25
C (characteristically).
Non-volatile memory consists of a block of user-allocated memory and a block of xed memory .
The user-allocated memory is available for you to save the following data:
power meter calibration data
user calibration kit de nitions
Even though calibration data is stored in non-volatile memory , if the associated
instrument state is not saved, you will not be able to retrieve the calibration
data after a power cycle .
The xed memory is used to store the following data (you cannot change where this data is
stored and it does not a ect your memory availability for storing user-allocated data):
copy con guration (printer and plotter type , port, baud rate , handshake)
power meter type (HP 436/438)
power sensor calibration factors and loss tables
user-de ned calibration kits
The maximum number of instrument states , calibrations , and memory traces that can reside
in non-volatile memory at any one time is limited to 31 instrument states , 128 calibrations (4
per instrument state , including the present instrument state), and 64 memory traces (2 per
instrument state , including the present instrument state).
In addition, the number of instrument states and associated calibrations and memory traces are
limited by the available memory . T o display the amount of unused memory on the analyzer ,
. (Be sure you have selected
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SA VE/RECALL
INTERNAL MEMORY as your disk type .) In the
upper right-hand portion of the display , the value displayed as Bytes free: is the unused
non-volatile memory . When you save to the internal memory , you will see the number of bytes
free decrease . When you delete les , the number of bytes free increases . There is a maximum
of 512 kBytes available .
If you have deleted registers since the last time the instrument was preset, the bytes available
for you to use may be less than the actual \bytes free" that is displayed. Deleting registers
to increase the available memory will work in cases where the registers being deleted and
the registers needing to be added are of the same standard size (such as instrument states not
having calibrations associated with them). In certain other cases , however , you may have
after deleting registers so that the \bytes free" value equals the available
memory value . During a preset, the analyzer runs a memory packer that de-fragments the free
memory into one contiguous block.
12-2 Preset State and Memory Allocation

T able 12-1 shows the memory requirements of calibration arrays and memory trace arrays
to help you approximate memory requirements . F or example , add the following memory
requirements:
a full 2-port calibration with 801 points (58 k)
the memory trace array (4.9 k)
the instrument state (3 k)
The total memory requirement is 65.9 kbytes . There is sucient memory to store 7 calibrations
of this type . However , the same calibration performed with 1601 points and 2 channels
uncoupled would require 252 k bytes:
a full 2-port calibration with 1601 points , two channels , uncoupled (230 k)
the memory trace array (19 k)
the instrument state (3 k)
Only 2 of these calibrations could reside in memory before the available memory would be
depleted.
T able 12-1.
Memory Requirements of Calibration and Memory Trace Arrays
V ariable Data Length (Bytes) Approximate T otals (Bytes)
401 pts 801 pts 1601 pts
1 chan 1 chan 2 chans
Response N 6 + 52 2.5 k 5 k 10 k 19 k
Response and isolation N 6 2 + 52 5 k 10 k 19 k 38 k
1-P ort N 6 3 + 52 7 k 14 k 29 k 58 k
2-P ort N 6 12 + 52 29 k 58 k 115 k 230 k
Interpolated cal Same as above in addition to regular cal
P ower Meter Cal
2 number of channels
) +208 1 k 1.8 k 3.4 k 6.6 k
N 6 + 52 2.5 k 4.9 k 9.7 k 19 k
3 k 3 k 3 k 3 k
N = number of points
This variable is allocated once per active channel.
The number of points that was set at the time the cal was turned on.
If the channels are coupled, this number is always 1. If the channels are uncoupled, this number
refers to the number of channels that have power meter cal on.
This value may change with di erent rmware revisions .
The analyzer attempts to allocate memory at the start of a calibration. If insucient memory
is available , an error message is displayed. It is possible that the CMOS memory might be
fragmented due to the sequence of saving and deleting states of various sizes . So another
alternative would be to store the current state to disk and then press
. The analyzer
runs a memory packer which might regain some previously inaccessible memory . If memory is
still inadequate , delete an instrument state and restart the calibration.
Preset State and Memory Allocation 12-3

Storing Data to Disk
Y ou can use the internal disk drive or connect an external disk drive for storage of instrument
states , calibration data, measurement data, and plot les . (Refer to Chapter 4, \Printing,
Plotting, and Saving Measurement Results", for more information on saving measurement data
and plot les .)
The analyzer displays one le name per stored instrument state when you list the disk
directory . In reality , several les are actually stored to the disk when you store the instrument
state . Thus , when the disk directory is accessed from a remote system controller , the directory
will show several les associated with a particular saved state . The maximum number of les
that you can store on a disk depends on the directory size . Y ou can de ne the directory size
when you format a disk. See T able 12-3 for the default directory size for oppy disks and hard
disks .
The maximum number of instrument states and calibrations that can reside on a disk is
limited by the available disk space . T o see the available disk space displayed on the analyzer ,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SA VE/RECALL
. (Be sure you have selected either
depending on your disk type .) In the upper righthand portion of the display , the value
displayed as Bytes free: is the available disk space . If your disk is formatted in LIF , this
value is the largest contiguous block of disk space . Since the analyzer is reporting the largest
contiguous block of disk space , you may or may not see the bytes free number change when
you delete les . If your disk is formatted in DOS, the number reported as bytes free is the total
available disk space . That number is updated whenever you save to or delete les from the
disk.
A disk le created by the analyzer appends a sux to the le name . (This is on the analyzer's
directory and is not visible .) The sux consists of one or two characters: the rst character is
the le type and the second is a data index. (Refer to T able 12-2 for the de nitions of each
sux character .)
T able 12-2. Sux Character De nitions
Char 1 De nition Char 2 De nition
G Graphics 1 Display graphics
D Error corrected data 1 Channel 1
R Raw data 1 to 4 Channel 1, raw arrays 1 to 4
5 to 8 Channel 2, raw arrays 5 to 8
F F ormatted data 1 Channel 1
C Cal K Cal kit
1 Cal data, channel 1 0 Stimulus state
1 to 9 Coecients 1 to 9
2 Cal data, channel 2 0 to C same as channel 1
M Memory trace data 1 Channel 1
If correction is on at the time of an external store , the calibration set is stored to disk. (Note
that inactive calibrations are not stored to disk.) When an instrument state is loaded into the
12-4 Preset State and Memory Allocation

analyzer from disk, the stimulus and response parameters are restored rst. If correction is on
for the loaded state , the analyzer will load a calibration set from disk that carries the same title
as the one stored for the instrument state .
If you are concerned about conserving memory , either internal memory or external disk space ,
some of the most memory-intensive operations include:
saving data arrays and graphics with the instrument state
Using Saved Calibration Sets
When you are saving to internal memory (CMOS, non-volatile memory), calibration sets are
linked to the instrument state and measurement parameter for which the calibration was
done . Therefore a saved calibration can be used for multiple instrument states as long as the
measurement parameter , frequency range , and number of points are the same . A full 2-port
calibration is valid for any S-parameter measurement with the same frequency range and
number of points . When an instrument state is deleted from memory , the associated calibration
set is also deleted if it is unused by any other state .
The following hints will help you avoid potential problems:
If a measurement is saved with calibration and interpolated calibration on, it will be restored
with interpolated calibration on.
A calibration stored from one instrument and recalled by a di erent one will be invalid. T o
ensure maximum accuracy , always recalibrate in these circumstances .
No record is kept in memory of the temperature when a calibration set was stored.
Instrument characteristics change as a function of temperature , and a calibration stored at
one temperature may be inaccurate if recalled and used at a di erent temperature . See
Chapter 7, \Speci cations and Measurement Uncertainties ," for allowable temperature ranges
for individual speci cations .
The HP 8753D can read disk les created by the HP 8753B/C and the HP 8753B/C can read
les created by the HP 8753D . A disk le translator is available to make HP 8753A disk les
compatible with HP 8753B les . These les can then be read by the HP 8753D . Contact your
local Hewlett-P ackard sales and service oce for a copy of this disk le translator .
Preset State and Memory Allocation 12-5

key is pressed, the analyzer reverts to a known state called the factory
preset state . This state is de ned in T able 12-3. There are subtle di erences between the
preset state and the power-up state . These di erences are documented in T able 12-4 . If power
to non-volatile memory is lost, the analyzer will have certain parameters set to default settings .
T able 12-5 shows the a ected parameters .
When line power is cycled, or the
key pressed, the analyzer performs a self-test
routine . Upon successful completion of that routine , the instrument state is set to the
conditions shown in T able 12-3 . The same conditions are true following a \PRES;" or \RST;"
command over HP-IB , although the self-test routines are not executed.
Y ou also can con gure an instrument state and de ne it as your user preset state:
PRESET
PRESET:USER .
key is now toggled to the
USER selection and your de ned instrument state will
be recalled each time you press
and when you turn power on. Y ou can toggle back to
N
NNNNNNNNNNNNNNNNNNNNNN
the factory preset instrument state by pressing
FACTORY .
When you send a preset over HP-IB , you will always get the factory preset.
Y ou can, however , activate the user-de ned preset over HP-IB by recalling the
register in which it is stored.
12-6 Preset State and Memory Allocation

T able 12-3. Preset Conditions (1 of 5)
Preset Conditions Preset V alue
Preset Conditions Preset V alue
Analyzer Mode Network Analyzer Mode
Frequency O set O
Edit Mode Start/Stop , Number of
O set V alue 0
Sweep Type Linear Frequency
F ormat Log Magnitude (all inputs)
Color Selections Same as before
Sweep Time 100 ms , A uto Mode
Start Frequency 30 kHz
A ctive Channel Channel 1
Frequency Span (1.) 2999.97 MHz
Frequency Span (Opt. 006) 5999.97 MHz
Intensity If set to 15%,
Time Span 100 ms
has no e ect. If set to
CW Frequency 1000 MHz
Source P ower 0 dBm
P ower Slope 0 dB/GHz; O
to 15%.
Start P ower 15.0 dBm
P ower Span 25 dB
Coupled P ower On
D2/D1 to D2 O
Source P ower On
Title Channel 1 = [hp]
Channel 2 = Empty
Coupled P ort P ower On
IF Bandwidth 3000 Hz
P ower Range A uto; Range 0
IF A veraging F actor 16; O
Number of P oints 201
Smoothing Aperture 1% SP AN; O
Phase O set 0 Degrees
Electrical Delay 0 ns
Preset State and Memory Allocation 12-7

T able 12-3. Preset Conditions (2 of 5)
Preset Conditions Preset V alue
Preset Conditions Preset V alue
P olar Marker Lin Mkr
Smith Marker R+jX Mkr
Calibration Kit 7 mm
System Z0 50 Ohms
V elocity F actor 1
P ort 1 0 s
Limit T esting O
P ort 2 0 s
Input A 0 s
Edit Mode Upper/Lower Limits
Input B 0 s
Stimulus O set 0 Hz
Chop A and B On
Amplitude O set 0 dB
P ower Meter Calibration O
Limit Type Sloping Line
Number of Readings 1
Beep F ail O
P ower Loss Correction O
Start Transform 20 nanoseconds
Transform Span 40 nanoseconds
Markers 1, 2, 3, 4, 5 1 GHz; All Markers O
Last A ctive Marker 1
Gate Start 10 nanoseconds
Gate Span 20 nanoseconds
Delta Marker Mode O
Marker T arget V alue 3 dB
Marker Width V alue 3 dB; O
HP-IB A ddresses Last A ctive State
HP-IB Mode Last A ctive State
Marker Stimulus O set 0 Hz
F ocus Last A ctive State
Marker V alue O set 0 dB
Clock Time Stamp On
Marker A ux O set (Phase) 0 Degrees
12-8 Preset State and Memory Allocation

T able 12-3. Preset Conditions (3 of 5)
Preset Conditions Preset V alue
Preset Conditions Preset V alue
P arallel P ort Last A ctive State
Spur A voidance On
Plotter Type Last A ctive State
A ux Input Resolution Low
Plotter P ort Last A ctive State
Analog Bus Node 11 (A ux Input)
Plotter Baud Rate Last A ctive State
Plotter Handshake Last A ctive State
HP-IB A ddress Last A ctive State
Printer Type Last A ctive State
Printer P ort Last A ctive State
Printer Baud Rate Last A ctive State
Plot T ext On
Printer Handshake Last A ctive State
Printer HP-IB A ddress Last A ctive State
Plot Quadrant Full P age
Disk Save Con guration
Plot Speed F ast
P en Number:
Raw Data Array O
F ormatted Data Array O
Select Disk Internal Memory
Ch1 T ext 7
Disk F ormat LIF
Ch2 T ext 7
Line Type:
Source Phase Lock Loop On
1 The directory size is calculated as 0.013% of the oppy disk size (which is 256) or 0.005% of the hard disk size .
2 Pressing preset turns o sequencing modify (edit) mode and stops any running sequence .
Preset State and Memory Allocation 12-9

T able 12-3. Preset Conditions (4 of 5)
Preset Conditions Preset V alue
Preset Conditions Preset V alue
Printer Mode Last A ctive State
A uto-F eed On
T able 12-3. Preset Conditions (5 of 5)
F ormat T able Scale
P osition V alue
Log Magnitude (dB) 10.0 5.0 0.0
Phase (degree) 90.0 5.0 0.0
Group Delay (ns) 10.0 5.0 0.0
Smith Chart 1.00 { 1.0
P olar 1.00 { 1.0
Linear Magnitude 0.1 0.0 0.0
Real 0.2 5.0 0.0
Imaginary 0.2 5.0 0.0
T able 12-4. P ower-on Conditions (versus Preset)
HP-IB MODE T alker/listener .
SA VE REGISTERS P ower meter calibration data and calibration data not associated with
an instrument state are cleared.
COLOR DISPLA Y Default color values .
INTENSITY F actory stored values . The factory values can be changed by running
the appropriate service routine . Refer to the \A djustments and
Correction Constants" chapter in the HP 8753D Service Guide.
SEQUENCES Sequence 1 through 5 are erased.
DISK DIRECTORY Cleared.
12-10 Preset State and Memory Allocation

T able 12-5. Results of P ower Loss to Non-V olatile Memory
HP-IB ADDRESSES are set to the following defaults:
HP 8753D : 16
USER DISPLA Y : 17
PLOTTER : 5
PRINTER : 1
POWER METER : 13
DISK : 0
DISK UNIT NUMBER : 0
DISK V OLUME NUMBER : 0
POWER METER TYPE is set to HP 438A/437
(store les) are set to defaults: FILE1 through FILE 5
PRINT TYPE is set to default: MONOCHROME
PRINTING/PLOTTING SETUPS are set to the following defaults:
P ARALLEL PORT : COPY
PLOTTER TYPE : PLOTTER
PLOTTER PORT : SERIAL
PLOTTER B A UD RA TE : 9600
PLOTTER HANDSHAKE : Xon-Xo
PRINTER TYPE : DESKJET
PRINTER PORT : P ARALLEL
PRINTER B A UD RA TE : 19200
PRINTER HANDSHAKE : Xon-Xo
1 Only applies to HP-IB operation.
Preset State and Memory Allocation 12-11



The CITI le Data F ormat and K eyword Reference
This appendix contains the following information:
The CITI le Data F ormat
De nition Of CITI le T erms
The CITI le Keyword Reference .
The CITI le Data F ormat
CITI le is a standardized data format, used for exchanging data between di erent computers
and instruments . CITI le is an abbreviation for \Common Instrumentation Transfer and
Interchange le". This standard has been a group e ort between instrument designers and
designers of computer-aided design programs . As much as possible , CITI le meets current needs
for data transfer , and it was designed to be expandable so it can meet future needs .
CITI le de nes how the data inside an ASCII package is formatted. Since it is not tied to any
particular disk or transfer format, it can be used with any operating system (B ASIC, DOS,
UNIX, etc .), with any disk format (LIF , DOS, 1, etc .), or with any transfer mechanism (disk,
LAN, GPIB , etc .).
By careful implementation of the standard, instruments and software packages using CITI le
are able to load and work with data created on another instrument or computer . It is possible ,
for example , for a network analyzer to directly load and display data measured on a scalar
analyzer , or for a software package running on a computer to read data measured on the
network analyzer .
There are two main types of data formats: binary and ASCII. CITI le uses the ASCII text
format. While this format does take up more bytes of space than a binary format, ASCII data
is a transportable , standard type of format which is supported by all operating systems . In
addition, the ASCII format is accepted by most text editors . This allows les to be created,
examined, and edited easily , making CITI le easier to test and debug.
File and Operating System F ormats
CITI le was designed to be independent of the data storage mechanism, and therefore may
be implemented for any le system. However transfer between le systems may sometimes
be necessary . Any commercially available software that has the ability to transfer ASCII les
between systems may be used to transfer CITI le data.
The CITI le Data Format and K eyword Reference A-1

De nition of CITI le T erms
This section will de ne the following terms:
A CITI le P ackage
A typical package is divided into two parts: The rst part, the header , is made up of keywords
and setup information. The second part, the data, usually consists of one or more arrays of
data. Example 1 shows the basic structure of a CITI le package:
Example 1, A CITI le P ackage
The \header" part CITIFILE A.01.00
V AR FREQ MA G 3
D A T A S RI
The \data" part BEGIN
0.23491E-3,-1.39883E-3
2.00382E-3,-1.40022E-3
END
When stored in a disk le there may be more than one CITI le package . With the 8510 network
analyzer , for example , storing a \memory all" will save all eight of the memories held in the
instrument. This results in a single le which contains eight CITI le packages .
The CITI le Header
The header section contains information about the data that will follow . It may also include
information about the setup of the instrument that measured the data. The CITI le header
shown in Example 1 has just the bare minimum of information necessary; no instrument setup
information was included.
An Array of Data
An array is numeric data that is arranged with one data element per line . A CITI le package
may contain more than one array of data. Arrays of data start after the BEGIN keyword, and
the END keyword will follow the last data element in an array .
A CITI le package does not necessarily need to include data arrays; for instance , CITI le could
be used to store the current state of an instrument. In that case the keywords V AR, D A T A,
BEGIN, and END would not be required.
A-2 The CITI le Data Format and K eyword Reference

CITI le K eyword
Keywords are always the rst word on a new line . They are always one continuous word
without embedded spaces .
A listing of all the keywords used in the latest A.01.01 version of CITI le is shown in \The
CITIFile Keyword Reference ."
When reading a CITI le , unrecognized keywords should be ignored. This allows new keywords
to be added, without a ecting an older program or instrument that might not use the new
keywords . The older instrument or program can still use the rest of the data in the CITI le as it
did before . Ignoring unknown keywords allows backwards compatibility to be maintained.
The CITI le Data Format and K eyword Reference A-3

Example 2, An 8510 Display Memory File
Example 2 shows a simple le that contains no frequency information. Some instruments do
not keep frequency information for display memory data, so this information is not included in
the CITI le package .
in this le . This convention allows the designer to de ne keywords that are particular to his or
her particular implementation.
#NA VERSION HP8510B.05.00
NAME MEMORY
#NA REGISTER 1
VAR FREQ MAG 5
DATA S RI
BEGIN
-1.31189E-3,-1.47980E-3
-3.67867E-3,-0.67782E-3
-3.43990E-3,0.58746E-3
-2.70664E-4,-9.76175E-4
0.65892E-4,-9.61571E-4
END
Example 3, 8510 Data le
Example 3 shows a CITI le package created from the data register of an 8510 Network
Analyzer . In this case 10 points of real and imaginary data was stored, and frequency
information was recorded in a segment list table .
#NA VERSION HP8510B.05.00
NAME DATA
#NA REGISTER 1
VAR FREQ MAG 10
DATA S[1,1] RI
SEG_LIST_BEGIN
SEG 1000000000 4000000000 10
SEG_LIST_END
BEGIN
0.86303E-1,-8.98651E-1
8.97491E-1,3.06915E-1
-4.96887E-1,7.87323E-1
-5.65338E-1,-7.05291E-1
8.94287E-1,-4.25537E-1
1.77551E-1,8.96606E-1
-9.35028E-1,-1.10504E-1
3.69079E-1,-9.13787E-1
7.80120E-1,5.37841E-1
-7.78350E-1,5.72082E-1
END
A-4 The CITI le Data Format and K eyword Reference
CITI le Examples · Example · Example

Example 4, 8510 3-T erm Frequency List Cal Set File
Example 4 shows how CITI le may be used to store instrument setup information. In the case
of an 8510 Cal Set, a limited instrument state is needed in order to return the instrument to the
same state that it was in when the calibration was done .
Three arrays of error correction data are de ned by using three D A T A statements . Some
instruments require these arrays to be in the proper order , from E1 to E3. In general, CITI le
implementations should strive to handle data arrays that are arranged in any order .
#NA VERSION HP8510B.05.00
NAME CAL_SET
#NA REGISTER 1
VAR FREQ MAG 4
DATA E[1] RI
DATA E[2] RI
DATA E[3] RI
#NA SWEEP_TIME 9.999987E-2
#NA POWER1 1.0E1
#NA POWER2 1.0E1
#NA PARAMS 2
#NA CAL_TYPE 3
#NA POWER_SLOPE 0.0E0
#NA SLOPE_MODE 0
#NA TRIM_SWEEP 0
#NA SWEEP_MODE 4
#NA LOWPASS_FLAG -1
#NA FREQ_INFO 1
#NA SPAN 1000000000 3000000000 4
#NA DUPLICATES 0
#NA ARB_SEG 1000000000 1000000000 1
#NA ARB_SEG 2000000000 3000000000 3
VAR_LIST_BEGIN
VAR_LIST_END
The CITI le Data Format and K eyword Reference A-5

1.12134E-3,1.73103E-3
4.23145E-3,-5.36775E-3
-0.56815E-3,5.32650E-3
-1.85942E-3,-4.07981E-3
END
BEGIN
2.03895E-2,-0.82674E-2
-4.21371E-2,-0.24871E-2
0.21038E-2,-3.06778E-2
1.20315E-2,5.99861E-2
END
BEGIN
4.45404E-1,4.31518E-1
8.34777E-1,-1.33056E-1
-7.09137E-1,5.58410E-1
4.84252E-1,-8.07098E-1
END
When an instrument's frequency list mode is used, as it was in Example 4, a list of frequencies
is stored in the le after the V AR LIST BEGIN statement. The unsorted frequency list
segments used by this instrument to create the V AR LIST BEGIN data are de ned in the #NA
ARB SEG statements .
The descriptions and examples shown here demonstrate how CITI le may be used to store and
transfer both measurement information and data. The use of a single , common format will
allow data to be more easily moved between instruments and computers .
A-6 The CITI le Data Format and K eyword Reference
Example (continued) · Conclusion

The CITI le K eyword Reference
K eyword Explanation and Examples
CITIFILE CITIFILE A.01.01 identi es the le as a CITI le , and indicates the
revision level of the le . The CITIFILE keyword and revision code must
precede any other keywords . The CITIFILE keyword at the beginning
of the package assures the device reading the le that the data that
follows is in the CITI le format. The revision number allows for future
extensions of the CITI le standard. The revision code shown here
following the CITIFILE keyword indicates that the machine writing this
le is using the A.01.01 version of CITI le as de ned here . Any future
extensions of CITI le will increment the revision code .
NAME NAME CAL_SET allows the current CITI le \package" to be named. The
name of the package should be a single word with no embedded spaces .
A list of standard package names follows:
RA W D A T A Uncorrected data.
D A T A Data that has been error corrected. When only a
single data array exists , it should be named D A T A.
FORMA TTED Corrected and formatted data.
MEMORY Data trace stored for comparison purposes .
CAL SET Coecients used for error correction.
CAL KIT Description of the standards used
DELA Y T ABLE Delay coecients for calibration.
V AR VAR FREQ MAG 201 de nes the name of the independent variable
(FREQ), the format of values in a V AR LIST BEGIN table (MA G, if
used), and the number of data points (201). Typical names for the
independent variable are FREQ (in Hz), TIME (in seconds), and POWER
(in dBm). F or the V AR LIST BEGIN table , only the \MA G" format is
supported at this point.
# #NA POWER1 1.0E1 allows variables speci c to a particular type of
device to be de ned. The pound sign (#) tells the device reading the
le that the following variable is for a particular device . The \NA"
shown here indicates that the information is for a Network Analyzer .
This convention allows new devices to be de ned without fear of
con ict with keywords for previously de ned devices . The device
identi er (i.e . NA) may be any number of characters .
SEG LIST BEGIN SEG_LIST_BEGIN indicates that a list of segments for the independent
variable follow . F ormat for the segments is: [segment type] [start]
[stop] [number of points]. The current implementation only
supports a single segment. If there is more than one segment, the
V AR LIST BEGIN construct is used. CITI le revision A.01.00 supports
only the SEG (linear segment) segment type .
SEG LIST END SEG_LIST_END de nes the end of a list of independent variable
segments .
The CITI le Data Format and K eyword Reference A-7
Label De nition

V AR LIST BEGIN VAR_LIST_BEGIN indicates that a list of the values for the independent
variable (declared in the V AR statement) follow . Only the MA G format
is supported in revision A.01.00.
V AR LIST END VAR_LIST_END de nes the end of a list of values for the independent
variable .
D A T A DATA S[1,1] RI de nes the name of an array of data that will be read
later in the current CITI le package , and the format that the data will
be in. Multiple arrays of data are supported by using standard array
indexing. V ersions A.01.00 and A.01.01 of CITI le only support the
RI (real and imaginary) format, and a maximum of two array indexes .
Commonly used array names include the following:
\V OL T A GE" Example: V OL T A GE[1]
\V OL T A GE RA TIO" for a ratio of two
voltages (A/R).
CONST ANT CONSTANT [name] [value] allows for the recording of values which
don't change when the independent variable changes .
CONST ANT s are part of the main CITI le de nition. Users must not
de ne their own CONST ANT s . Use the #KEYWORD device speci cation
to create your own KEYWORD instead. The #NA device speci cation is
an example of this .
No constants were de ned for revision A.01.00 of CITI le . CITI le
revision A.01.01 de ned the following constant:
CONST ANT TIME [year] [month] [day] [hour] [min] [secs]
Example:
COMMENT YEAR MONTH D A Y HOUR MINUTE SECONDS
CONST ANT TIME 1991 02 26 17 33 53.25
The COMMENT statement is not absolutely required, but is highly
recommended to aid readability .
The year should always be the full four digits (\1991" is correct, but
\91" is not). This is to avoid problems with the year 2000, when the
shortened version of the year will be \00."
The hour value should be in 24-hour \military" time .
When writing a CITI le and the fractional seconds value is zero , then
the \seconds" value may be printed either with or without a decimal
a CITI le , the seconds value should always be read as if it were a
oating point number .
A-8 The CITI le Data Format and K eyword Reference

battery backup for memory, 12-2
measurement setup diagram, 3-31
bias inputs and fuses locations, 1-11
annotations of display, 1-7
blanking the display, 6-42
application and operation concepts, 6-1
broad band power meter calibration
ampli er testing, 6-145{148
C10 (pass control capabilities), 11-18
C1,C2,C3 (system controller capabilities),
raw data , 4-36
assign classes, 5-29, 5-31
C2 status notation, 1-7
A TN (attention) control line, 11-17
attenuation at mixer ports, 3-2
allowing repetitive switching, 6-19
test port return, 11-2
AUTO-FEED , 4-12
calculate statistics of the measurement data,
automatic le naming convention for LIF
auto sweep time mode
how to set, 5-51
auxiliary input connector location, 1-11
for mixer measurements, 3-6
for non-insertable devices, 6-101
for noninsertable devices, 5-41
how to widen, 5-52
A VG
menu map, 8-2
A vg status notation, 1-7
power meter, 5-33, 6-95
fault location measurements, 6-123
procedure to modify a user kit, 6-76
re ection measurements, 6-119
response and isolation, 6-73
system:how to widen, 5-52
basic measurement sequence and example,
basic system maintenance service option,
basic talker (T6), 11-18

channel display titling, 2-8
using on multiple analyzers, 12-5
verify performance of, 6-84
channel stimulus coupling, 6-20
characteristic power meter calibration speed
saving modi ed kit, 5-28
characterizing microwave systematic errors,
calibration factors entry, 5-34
characters that are valid, 11-22
label standard menu, 6-81
chop sweep mode, 6-23, 6-100
saving modi ed or user kit, 5-28
how to activate, 5-54
specify o set, 6-80
class label, 5-30, 5-32
standard de nitions, 6-78
cleaning of connectors, 2-2
clearing a sequence from memory, 2-57
clock setting, 4-30, 11-13
code naming conventions, 11-21
color printer use, 4-6
electrical delay of, 5-6
electrical o set, 6-67
frequency response of, 6-66
to key cross reference, 9-48
outline of modi cation, 5-26
cal kit menu, 6-75
menu map, 8-3
directional coupler response, 5-35
capabilities of the analyzer, 1-2
electrical delay of calibration standards,
care of connectors, 2-2
cascading multiple sequences, 2-61
complete service request capabilities (SR1),
completion time for measurements, 7-33
center frequency set with markers, 2-18
centronics (parallel) interface, 1-10
what you can save to , 4-34
changes between the HP 8753A/B/C/D, 1-13
change system bandwidth, 5-57
application and operation, 6-1
changing the sequence title, 2-58
channel 1 and 2 ratio measurement, 2-7

care quick reference, 2-2
for external monitor, 1-10 , 1-11
parallel (centronics) interface, 1-10
probe power source, 1-5
serial (RS-232) interface, 1-10
conditions for environment, 7-36
conditions for error-correction, 5-4
test set interconnect, 1-11
external disk drive, 11-12
plotter, 4-8 , 11-9
connectors on front panel, 7-33
connectors on rear panel, 7-34
a plot function, 4-8
considerations for mixer measurements, 3-2
a print function, 4-3
of disk le, 4-33
printer, 4-4 , 4-9, 11-8
of internal memory registers, 4-33
the analyzer for the printer, 4-4
continuous and discrete markers, 2-9
con guring the analyzer for the peripheral,
continuous correction mode for power meter
continuous mode , test set switch, 5-54
broad band power meter calibration, 3-13
conversion loss using the tuned receiver
controlling the test set switch, 5-54
down converter port, 3-3
conventions for code naming, 11-21
rst portion of conversion compression
conversion compression mixer measurement,
group delay measurement, 3-25
and output power as a function of input
high dynamic range swept IF conversion
power level example, 3-28
mixer isolation measurement, 3-34
mixer RF feedthrough measurement, 3-36
one-sweep power meter calibration for
using the frequency o set mode, 3-7
using the tuned receiver mode connections,
peripheral to analyzer, 4-8
printer to analyzer, 4-3
using tuned receiver mode, 3-17
Rchannel and source calibration, 3-8
menu map, 8-5
response calibration, 3-18, 3-34, 3-36
correcting for measurement errors, 6-71
second portion of conversion compression
up converter port, 3-3

Cor status notation, 1-7
plotter pen numbers, 4-13
couple and uncouple display markers, 2-14
de ned plotting components, 4-12
de ne standard menus, 6-78
de nition of front panel keys, 9-2
creating a sequence, 2-54
cross reference of key function to HP-IB
all instrument state les, 4-41
uncorrected speci cations, 7-2
an instrument state, 4-41
C? status notation, 1-7
a single instrument state le, 4-41
CW time sweep, 6-23
CW time-to-frequency domain, 6-134
Del status notation, 1-7
delta () markers, 2-11
description and options, 1-1
description of hardkeys and softkeys, 9-2
divided by memory, 2-7
loss of power calibration, 5-33
deviation from linear phase measurement,
device clear (DC1), 11-18
device types for HP-IB, 11-15
trace saved to the display memory, 2-6
diagram of measurement frequencies, 3-10
port extensions and electrical delay, 5-3
data and memory viewing, 2-6
the HP 8753A/B/C/D, 1-13
dimensions of 7 mm test port, 7-33
dimensions of analyzer, 7-37
de nition of, 6-5
directional coupler compensation, 5-35
data processing ow diagram, 4-37
uncorrected speci cations, 7-2
for internal memory, 4-33
discrete and continuous markers, 2-9
DC1 (complete device clear), 11-18
the sweep rate, 5-48
storage of a sequence, 2-59
the time delay, 5-49
line types for plotter, 4-14
con guring to plot, 4-11
pen numbers for plot elements, 4-13
using with other models, 12-5
dispersion e ects, 6-94

E2 (tri-state drivers), 11-18
displayed measurement title, 4-29
of calibration standards, 5-6
electrical length and phase distortion
causes of measurement problems, 5-48
how to improve measurement results, 5-48
making accurate measurements of, 5-48
electrical o set, 6-67
end or identify (EOI) control line, 11-17
display functions, 2-5, 6-39
ensuring valid calibration, 6-71
entering calibration constants, 5-26
entry block location, 1-5
DISPLA Y
menu map, 8-6
divide measurement data by the memory
EOI (end or identify) control line, 11-17
error-correction, 2-4, 5-2, 5-7
does not respond to parallel poll (PPO), 11-18
characteristics a ected, 5-4
down converter port connections, 3-3
full two-port calibration, 6-73
full two-port measurements, 5-20
DT1 (responds to a group execute trigger),
dual channel mode, 2-5, 6-40
one-port re ection measurements, 5-17
response and isolation, 6-73
response and isolation for re ection
duration of measurements, 7-33
response and isolation for transmission
response for re ection measurements, 5-8
response for transmission measurements,
TRL*/LRM* two-port calibration, 6-74

CW frequency range, 6-111
external trigger circuit, 7-35
external trigger (EXT TRIGGER), 7-34
external video monitor, 7-35
ext mon. connector, 1-10, 1-11
examining calibration constants, 5-26
ext status notation, 1-7
conversion loss and output power as a
function of input power level, 3-28
conversion loss measurement, 3-11
fast 2-port calibration, 5-54
xed IF mixer measurement, 3-23
faster sweep speed, 5-50
group delay measurement, 3-26
mixer LO to RF isolation measurement,
using low pass mode, 6-123
feature di erences between
mixer RF feedthrough measurement, 3-37
plot quadrants, 4-17, 4-28
plotting two les on the same page, 4-27
printing two measurements, 4-7
features added to the analyzer, 1-2
signal ow in a mixer, 3-33
features of front panel, 1-4
spectrum of RF ,LO ,and IF signals present
in a conversion loss measurement, 3-7
swept IF conversion loss measurement,
swept power conversion compression
how to re-save, 4-41
viewing both channels with a split display,
extended listener capabilities (LEO), 11-18
internal memory registers, 4-33
am connector location, 1-11
trigger connector location, 1-11
ltering for mixers, 3-2
external AM input (EXT AM), 7-34
rst portion of conversion compression
external auxiliary input (A UX INPUT), 7-34
xed IF mixer measurement
xed IF mixer measurements, 3-17
how to solve problems with, 4-43
ag markers for gating, 2-70
external monitor connector, 1-10
at limit lines, 2-37
external reference frequency input (EXT
oppy disk le contents, 4-33
external source mode, 6-109
area of display, 1-8

F ORMA T
menu map, 8-6
gain and reverse isolation measurement,
FORMFEED , 4-29
gain compression measurement, 2-45
demodulating the results, 6-135
gate span setting, 2-70
response and isolation error-corrections,
selection for mixers, 3-2
gating e ects in frequency domain, 2-72
Gat status notation, 1-7
transform from CW time, 6-134
GPIO mode, 6-105, 6-139
frequency o set mode, 3-7
frequency o set operation, 6-112
compatible instrument modes, 6-113
group delay accuracy vs . aperture, 7-31
group delay or phase mixer measurements,
group execute trigger response (DT1), 11-18
guidelines for code naming, 11-21
o set frequency, 6-113
receiver and source requirements, 6-114
H=2 status notation, 1-8
H=3 status notation, 1-8
hardkey de nitions, 9-2
harmonic mode option, 1-12
coupling channel power, 6-115
frequency response of calibration standards,
front panel access key for softkeys, 9-69
front panel connectors, 7-33
front panel features, 1-4
front-panel key de nitions, 9-2
high dynamic range swept IF conversion loss
full-acceptor handshake (AH1), 11-18
full-source handshake (SH1), 11-18

high dynamic range swept RF/IF conversion
error-correct for frequency response, 5-8
error-correct for frequency response and
high-stability frequency reference(10 MHz)
error-correct for full two-port
high stability frequency reference option,
error-correct for one-port re ection
Hld status notation, 1-8
hold mode , test set switch, 5-54
error-correct for response and isolation
for re ection measurements, 5-13
abort a print or plot process, 4-30
error-correct for response and isolation
activate a xed marker, 2-11
for transmission measurements, 5-15
activate chop sweep mode, 5-54
error-correct for TRL*, 5-23
activate display markers, 2-10
error-correct for TRM*, 5-23
calculate data statistics, 2-26
calibrate the analyzer receiver to measure
error-correct response for re ection
change measurement averaging, 5-56
error-correct response for transmission
change system bandwidth, 5-56
change the sequence title, 2-58
format a disk, 4-43
gate out responses, 2-70
clear a sequence from memory, 2-57
increase sweep speed, 5-50
compensate for directional coupler
insert a sequence command, 2-56
load a sequence from disk, 2-60
con gure a plot function, 4-8
make a basic measurement, 2-3
con gure a plot function to disk drive,
make accurate measurements of electrically
con gure a plot function using and HPGL/2
measure conversion loss using tuned
con gure a plot function using a pen
measure deviation from linear phase, 2-32
measure devices using the time domain,
con gure a print function, 4-3
construct a loop structure in a sequence,
measure electrical length and phase
control the test set switch, 5-54
measure xed IF mixers, 3-17
couple and uncouple display markers, 2-14
measure gain compression, 2-45
create a limit test sequence, 2-64
measure group delay, 2-33
create a test sequence, 2-54
measure group delay or phase for mixers,
create at limit lines, 2-37
create single point limits, 2-41
measure high dynamic range conversion
create sloping limit lines, 2-39
decrease frequency span, 5-50
measure magnitude and insertion phase
de ne line types, 4-14
de ne the plot, 4-12
measure mixer isolation, 3-33
de ne the print, 4-5
measure mixer LO to RF isolation, 3-33
delete all instrument state les, 4-41
delete an instrument state, 4-41
measure re ection response in time domain,
delete a single instrument state le, 4-41
delete limit segments, 2-42
measure reverse isolation and gain, 2-49
divide measurement data by the memory
measure swept harmonics, 2-66
measure swept RF/IF mixers, 3-7
edit a sequence, 2-56
measure transmission response in time
edit limit segments, 2-42
edit power calibration factors, 5-34
modify a sequence command, 2-57
enter the power sensor calibration data,
modify calibration standards, 5-26
modify TRL calibration standards, 5-28

modify TRM calibration standards, 5-30
o set limit lines, 2-44
set auto sweep time mode, 5-51
output a single page of values, 4-30
set center frequency with markers, 2-18
output multiple plots to a single page using
set CW frequency using markers, 2-21
set display reference using markers, 2-20
output plot les, 4-20
set electrical delay using markers, 2-21
output plot les from a PC to a plotter,
set frequency range, 2-3
set frequency span with markers, 2-19
output plot les from a PC to a printer,
set source power, 2-4
set start frequency with markers, 2-17
output single page plots using a printer,
set stop frequency with markers, 2-17
set the measurement parameters, 2-3
output the entire list of values, 4-31
set the measurement type, 2-4
plot a measurement to disk, 4-19
set the sweep type, 5-53
plot measurements in page quadrants,
solve printing or plotting problems, 4-32
solve problems with disk storage, 4-43
plot multiple measurements on a full page,
stop a sequence, 2-55
store a sequence on a disk, 2-59
plot multiple measurements per page,
subtract the memory trace from the
measurement data trace, 2-7
plot one measurement per page, 4-16
title the active channel display, 2-8
plot to an HPGL compatible printer, 4-18
title the displayed measurement, 4-29
power meter calibrate, 3-6
print a sequence, 2-60
use analyzer display markers, 2-9
print multiple measurements per page,
use continuous and discrete markers, 2-9
use delta () markers, 2-11
print one measurement per page, 4-6
use frequency o set mode, 3-7
print or plot list values or operating
use limit lines, 2-36
MKR ZERO to activate a xed reference
purge a sequence from disk, 2-60
ratio measurements in channel 1 and 2,
use polar format markers, 2-14
use receiver calibration, 5-11
recall a le, 4-42
use Smith chart markers, 2-15
reduce receiver crosstalk, 5-57
use the tuned receiver mode, 2-51
reduce receiver noise oor, 5-56
REF=FIXED MKR key to activate
reduce the averaging factor, 5-52
a xed reference marker, 2-12
reduce the number of points, 5-52
view a single measurement channel, 5-53
reduce trace noise, 5-57
view both measurement channels, 2-5
rename a le, 4-42
view plot les on a PC, 4-20
re-save a le, 4-41
view the measurement data and memory
reset plotting parameters to default values,
widen system bandwidth, 5-52
reset the printing parameters to default
HP 8753A/B/C/D di erences, 1-13
review limit line segments, 2-43
run a limit test, 2-43
outputting plot les to compatible printer
run a test sequence, 2-55
save a data trace to the display memory,
test le commands, 4-24
HPGL/2 compatible printer , con guring to
save an instrument state, 4-35
save measurement results, 4-36
search for a target amplitude, 2-24
search for bandwidth, 2-25
search for maximum amplitude, 2-22
bus structure, 11-15, 11-16
search for minimum amplitude, 2-23

instrument state function block location, 1-5
instrument state in memory, 12-1
instrument states in memory, 12-2
interconnect devices ,max.number, 11-17
message transfer scheme, 11-17
interconnection cable ,max. length, 11-17
pass control mode, 6-104
status indicators, 6-104, 11-19
system controller mode, 6-104
interface clear (IFC) control line, 11-16
HP-IB-command to key-function cross
interface function codes, 7-33
how de ned for mixers, 3-2
how to widen, 5-52
IFC (interface clear) control line, 11-16
interpolated error-correction, 5-6, 6-72
measurements for mixers, 3-33
(reverse) and gain measurement, 2-49
test port power, 5-56
increasing measurement accuracy, 5-2
key de nitions, 9-1
key menu maps, 8-1
LIF, 11-5
input ports menu, 6-28
inserting a sequence command, 2-56
insertion phase and magnitude response
DISPLA Y
frequency o set, 6-112
F ORMA T
LOCAL
, 6-26, 6-27
MENU
deleting all les, 4-41
deleting a single le, 4-41

line types available, 4-14
line voltage selector switch, 1-10
LIN MKR, 2-15, 2-16
listener interface function, 11-15
listen mode (L), 11-19
keys referenced to front panel access key,
key to HP-IB command, 9-48
list frequency sweep, 6-22
kit label, 5-30, 5-32
printing or plotting, 4-30
L (listen mode), 11-19
loading a sequence from disk, 2-60
label class menu, 6-83
uncorrected speci cations, 7-2
label kit, 5-30, 5-32
LOCAL
menu map, 8-7
label the classes, 5-30, 5-32
10 MHz precision reference output, 1-11
LE0 (no extended listener capabilities), 11-18
10 MHz reference adjust, 1-11
active channel keys, 1-4
auxiliary input connector, 1-11
bias inputs and fuses, 1-11
centronics (parallel) interface, 1-10
external trigger connector, 1-11
instrument state function block, 1-5
limit test connector, 1-11
creating at limit lines, 2-37
creating single point limits, 2-41
line voltage selector switch, 1-10
creating sloping limit lines, 2-39
P1 and P2 in
SCALE PLOT [GRAT] mode,
deleting limit segments, 2-42
editing limit segments, 2-42
parallel (centronics) interface, 1-10
o setting limit lines, 2-44
port 1 and port 2, 1-5
reviewing limit line segments, 2-43
running a limit test, 2-43
probe power source connector, 1-5
R channel connectors, 1-5
limit test output (LIMIT TEST), 7-35
response function block, 1-4
RS-232 (serial) interface, 1-10
phase deviation measurement, 2-32
serial number plate, 1-10
linear frequency sweep, 6-21
serial (RS-232) interface, 1-10
linear magnitude format, 6-33
lines for control, 11-16
stimulus function block, 1-4
lines for handshaking, 11-16
test sequence connector, 1-11
line switch location, 1-4
test set interconnect, 1-11
LINE TYPE DATA , 4-14
locations of softkeys, 9-69
logarithmic frequency sweep, 6-22
log magnitude format, 6-29
loop counter sequence, 2-62

loss of power meter calibration data, 5-33
setting center frequency, 2-18
LO to RF isolation for mixers, 3-33
setting start frequency, 2-17
setting stop frequency, 2-17
re ection measurements, 6-123
setting frequency range, 6-122
Smith chart markers, 2-15
Smith or polar format, 2-12
magnitude and insertion phase response
bandwidth on display, 1-9
electrically long devices, 5-48
man status notation, 1-8
using to calibrate non-insertable devices,
values area on display, 1-8
using to calibrate noninsertable devices,
menu map, 8-8
menu map, 8-8
math with data traces, 2-7
maximum amplitude search, 2-22
key, 6-26, 6-27
active and inactive, 2-10
menu map, 8-9
calculate data statistics, 2-26
continuous and discrete, 2-9
deviation from linear phase, 2-32
couple and uncouple display markers, 2-14
device under test, 2-4
electrical length and phase distortion, 2-30
gain and reverse isolation, 2-49
magnitude and insertion phase response,
functions on display, 2-9
transmission response in time domain,
using limit lines, 2-36
using the time domain, 2-68
polar format, 2-14, 6-49
measured inputs of display, 1-8
Re/Im MKR, 2-15, 2-16
conversion loss example, 3-11
data points:setting, 2-4
search for bandwidth, 2-25
search for maximum amplitude, 2-22
search for minimum amplitude, 2-23
search for target amplitude, 2-24
parameters set with markers, 2-16
set CW frequency, 2-21
set display reference, 2-20
type:setting, 2-4
set electrical delay, 2-21
set frequency span, 2-19
set measurement parameters, 2-16

compensating for directional coupler
using the tuned receiver mode, 2-51
measurement setup diagram shown on
measurement setup from display, 3-10
performance veri cation, 5-3
3.5 mm test ports, 7-8
reference plane and port extensions, 5-3
75 type-F test ports, 7-14, 7-17, 7-20,
75 type-N test ports, 7-11
7 mm test ports, 7-2
measurement channel, how to view a single,
saving calibration data, 12-2
saving current instrument state, 12-1
measurement frequencies , diagram, 3-10
saving instrument states, 12-2
how to reduce, 5-52
measurement port speci cations, 7-2
uncorrected for errors, 7-2
memory math functions, 6-41
electrically long devices, 5-48
conversion loss for mixers, 3-7
conversion loss using tuned receiver mode,
high dynamic range conversion loss for
LO to RF isolation for mixers, 3-33
mixer conversion compression, 3-28
mixer group delay or phase, 3-24
mixers xed IF, 3-17
A VG
CAL
power meter calibration for, 3-6
RF feedthrough for mixers, 3-35
swept RF/IF mixers, 3-7

MARKER
MARKER F CTN
amplitude and phase tracking, 6-159
MENU
PRESET
SA VE/RECALL
SCALE REF
SEQ
SYSTEM
frequency o set mode, 6-149
message area of display, 1-8
message transfer scheme, 11-17
methods of HP-IB operation, 11-15
microwave connector care, 2-2
minimum loss pad, 11-2
tuned receiver mode, 6-149
isolation measurement connections, 3-34
MKR ZERO example procedure, 2-13
LO to RE isolation measurement example,
auto sweep time, 5-51, 6-16
RF feedthrough measurement connections,
frequency o set, 3-7
manual sweep time, 6-16
RF feedthrough measurement example,
tuned receiver , 3-17
amplitude and phase tracking, 3-27
frequency o set, 6-112
attenuation at ports, 3-2
conversion loss using frequency o set, 3-7
conversion loss using tuned receiver mode,
eliminating unwanted signals, 3-2
modes for bus device, 11-19
frequency o set mode operation, 3-4
modi ed user kit, 5-28
group delay or phase measurement, 3-24
calibration kit menu, 6-77
high dynamic range conversion loss, 3-12
the cal kit thru de nition, 5-40
how RF and IF are de ned, 3-2
LO to RF isolation, 3-33
a sequence command, 2-57
minimizing source and load mismatches,
power meter calibration, 3-6
reducing the e ect of spurious responses,
TRL calibration standards, 5-28
TRM calibration standards, 5-30
swept RF/IF measurement, 3-7

of the network analyzer, 6-2
operational capabilities for HP-IB, 11-18
operation of HP-IB, 11-15
multiple sequence cascading, 2-61
network analyzer mode, 6-109
new features to the analyzer, 1-2
002, harmonic mode, 1-12
NEWLINE , 4-29
006, 6 GHz operation, 1-12
no extended talker capabilities (TEO), 11-18
010, time domain, 1-12
011, receiver con guration, 1-12
+02A, standard system maintenance
noise level speci cations, 7-28
+02B , basic system maintenance service,
075, 75 impedance, 1-12
1CM, rack mount ange kit without handles,
1CP , rack mount ange kit with handles,
1D5, high stability frequency reference,
non-operating storage conditions, 7-36
+22A, return to HP full service agreement,
22B , return to HP repair agreement, 1-13
+22C, return to HP calibration agreement,
number of HP-IB devices allowed, 11-15
number of listeners allowed, 11-15
+22G, return to HP calibration, 1-13
+23G, on-site system veri cation, 1-13
how to reduce, 5-52
service and support, 1-13
number of sweeps , test set switch, 5-54
options and description, 1-1
o setting limit lines, 2-44
Ofs status notation, 1-7
Of? status notation, 1-7
one-sweep power meter calibration for mixer
multiple plots to a single page using a
on-site system veri cation option, 1-13
single page plots using a printer, 4-24
from a PC to a plotter, 4-22
printing or plotting, 4-30
from a PC to a printer, 4-23
frequency o set, 6-112

P1 and P2 on the plotter, 4-15
con gurations, 4-4, 4-8, 11-8
connections to analyzer, 4-8
page of values printed or plotted, 4-30
parallel poll non response (PPO), 11-18
phase and magnitude response measurement,
GPIO mode, 6-105, 6-139
phase distortion and electrical length
de ned boundaries, 4-15
defaults for plotting, 4-16
defaults for printing, 4-6
operating:printing or plotting, 4-30
multiple measurements per page, 4-26
parameters set with markers, 2-16
PLOT DATA , 4-12
how to view on a PC, 4-20
outputting from a PC to a plotter, 4-22
outputting from a PC to a printer, 4-23
PLOT GRAT , 4-12
NNNNNNNNNNNNNNNNNNNNNNNNNN
PLOT MEM , 4-12
NNNNNNNNNNNNNNNNNNNNNNNNNN
PLOT MKR , 4-12
pass control capabilities (C10), 11-18
con guration, 4-8, 11-9
indicators in limit testing, 2-44
pen number settings, 4-13
pass fail indicators on display, 1-9
plotter P1 and P2, 4-15
PC? status notation, 1-8
PC status notation, 1-8
P# status notation, 1-8
pen number settings, 4-13
a measurement to disk, 4-19
PEN NUM DATA , 4-13
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
components de ned, 4-12
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PEN NUM MARKER , 4-13
multiple measurements on a full page,
PEN NUM MEMORY , 4-13
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
multiple measurements per page, 4-17
con guring to plot, 4-10, 11-11
to an HPGL compatible printer, 4-18
plotting multiple measurements per page,
plotting, printing, and saving, 4-1
point limit creation, 2-41

PPO (does not respond to parallel poll, 11-18
data:how to reduce, 5-52
pre-raw data arrays, 6-6
LOG MKR, 2-15
PRESET
menu map, 8-12
principles of microwave connector care, 2-2
polar or Smith format markers, 2-12
port 1 and port2, 1-5
multiple measurements per page, 4-7
PRINT ALL use, 4-31
increasing test port, 5-56
con guration, 4-4, 4-9, 11-8
outputting single page plots to, 4-24
using to output multiple plots to a single
power cord receptacle with fuse, 1-10
power level, calibrated, 6-95
printer connections to analyzer, 4-3
calibration for mixer measurements, 3-6
power meter calibration, 6-95
printing a sequence, 2-60
compatible sweep types, 6-95
printing or plotting list values or operating
continuous correction mode, 5-37
continuous sample mode, 6-96
printing, plotting, and saving, 4-1
print/plot time stamp, 4-30, 11-13
PRm status notation, 1-8
loss of calibration data, 5-33
modes of operation, 6-96
probe power source connector location, 1-5
power loss correction list, 6-98
with disk storage, 4-43
power sensor calibration data, 6-98
with printing or plotting, 4-32
receiver to measure absolute power, 5-38
sample-and-sweep correction mode, 5-36
aborting a print or plot process, 4-30
activating a xed marker, 2-11
activating display markers, 2-10
basic measurement sequence, 2-3
power (output) characteristics, 7-27
calculating data statistics, 2-26
calibrating the analyzer receiver to measure
changing the sequence title, 2-58
clearing a sequencing from memory, 2-57
compensating for directional coupler
calibration factor list, 6-98
con guring a plot function, 4-8
con guring a print function, 4-3

constructing a loop structure in a sequence,
modifying TRMM calibration standards,
coupling and uncoupling display markers,
o setting limit lines, 2-44
outputting a single page of values, 4-30
creating a sequence, 2-54
outputting multiple plots to a single page
creating at limit lines, 2-37
using a printer, 4-25
creating single point limits, 2-41
outputting plot les, 4-20
creating sloping limit lines, 2-39
outputting plot les from a PC to a plotter,
de ning line types, 4-14
de ning the plot, 4-12
outputting plot les from a PC to a printer,
de ning the print, 4-5
deleting an instrument state, 4-41
outputting single page plots using a printer,
deleting a single instrument state le, 4-41
deleting limiting segments, 2-42
outputting the entire list of values, 4-31
deviation from linear phase measurement,
plotting a measurement to disk, 4-19
plotting measurements in page quadrants,
dividing measurement data by the memory
plotting multiple measurements on a full
editing a sequence, 2-56
editing limit segments, 2-42
plotting multiple measurements per page,
entering the power sensor calibration data,
plotting to an HPGL compatible printer,
error-correction for full two-port
printing a sequence, 2-60
error-correction for one-port re ection
printing multiple measurements per page,
error-correction for TRL*, 5-23
printing one measurement per page, 4-6,
error-correction for TRM*, 5-23
formatting a disk, 4-43
printing or plotting list values or operating
frequency response error-corrections, 5-8
gating in time domain, 2-70
purging a sequence from disk, 2-60
inserting a sequence command, 2-56
ratioing measurements in channel 1 and
limit testing using a sequence, 2-64
loading a sequence from disk, 2-60
recalling a le, 4-42
making accurate measurements of
renaming an instrument state le, 4-42
electrically long devices, 5-48
re-saving an instrument state, 4-41
measuring devices using the time domain,
resetting plotting parameters to default
measuring electrical length and phase
resetting the printing parameters to default
measuring gain compression, 2-45
response and isolation error-correction for
measuring group delay, 2-33
re ection measurements, 5-13
measuring magnitude and insertion phase
response and isolation error-correction for
measuring re ection response in time
response error-correction for re ection
measuring reverse isolation and gain, 2-49
response error-correction for transmission
measuring swept harmonics, 2-66
measuring transmission response in time
reviewing limit line segments, 2-43
running a limit test, 2-43
mixer group delay or phase, 3-24
running a sequence, 2-55
modifying a sequence command, 2-57
saving a data trace to the display memory,
modifying calibration standards, 5-26
modifying TRL calibration standards, 5-28
saving an instrument state, 4-35
searching for a target amplitude, 2-24

searching for bandwidth, 2-25
programming speci cations, 7-33
searching for maximum amplitude, 2-22
searching for minimum amplitude, 2-23
P? status notation, 1-8
setting center frequency with markers,
purging a sequence from disk, 2-60
purpose and use of di erent error-correction
setting CW frequency using markers, 2-21
setting display reference using markers,
setting electrical delay using markers, 2-21
setting frequency span with markers, 2-19
quarter page plots, 4-28
setting measurement parameters, 2-3
setting start frequency with markers, 2-17
setting stop frequency with markers, 2-17
rack mount cabinet, 11-4
setting the frequency range, 2-3
rack mount ange kit
setting the measurement type, 2-4
with handles option, 1-12
setting the source power, 2-4
without handles option, 1-12
setting up a color printer, 4-6
solving problems with disk storage, 4-43
dynamic:how to increasing, 5-56
stopping a sequence, 2-55
storing a sequence on a disk, 2-59
subtracting memory trace from
ratio measurement in channel 1 and 2, 2-7
measurement data trace, 2-7
raw data arrays, 6-6
R channel and source calibration connections,
test with limit lines, 2-36
titling the active channel display, 2-8
R channel connector locations, 1-5
titling the displayed measurement, 4-29
R-Channel inputs , di erence between internal
using analyzer display markers, 2-9
R-Channel power o set, 3-4
using continuous and discrete markers,
using delta () markers, 2-11
features and connectors, 1-10
MKR ZERO to activate a xed
rear panel connectors, 7-34
recalling a le, 4-42
using polar format markers, 2-14
using the tuned receiver mode, 2-51
REF=FIXED MKR key to activate
a xed reference marker, 2-12
receiver calibration procedure, 5-38
viewing both measurement channels, 2-5
receiver con guration option, 1-12
viewing plot les on a PC, 4-20
viewing the measurement data and memory
LO to RF isolation for mixers, 3-33
mixer conversion compression, 3-28
number of measurement points, 5-52
mixer RF feedthrough, 3-35
receiver noise oor, 5-56
power meter calibration for mixer
programming-command to key-function cross

(10 MHz) adjust, 1-11
return to HP calibration agreement option,
(10 MHz) output, 1-11
level of display, 1-8
return to HP calibration option, 1-13
return to HP full service agreement option,
return to HP repair agreement option, 1-13
reverse isolation and gain measurement,
using bandpass mode, 6-119
reviewing the limit line segments, 2-43
using low pass mode, 6-123
re ection response in time domain, 2-73
how de ned for mixers, 3-2
uncorrected speci cations, 7-2
RL1 (complete remote/local capability), 11-18
R (remote operation), 11-19
RS-232 (serial) interface, 1-10
Re/Im MKR, 2-15, 2-16
rules for code naming, 11-21
a limit test, 2-43
relative velocity for time domain, 2-75
sample-and-sweep correction mode, 5-36
remote enable (REN) control line, 11-17
remote/local capability (RL1), 11-18
for power meter calibration, 5-36
remote operation (R), 11-19
remote programming speci cations, 7-33
save a data trace to the display memory, 2-6
renaming a le, 4-42
menu map, 8-11
REN (remote enable) control line, 11-17
a user kit, 5-28
places where you can, 4-33
to a computer, 4-34
re-save a le, 4-41
user kit to disk, 5-28
saving measurement results including
plotting parameters to default values, 4-16
saving, printing, and plotting, 4-1
scale/div . area of display, 1-8
SCALE PLOT , 4-15
calibration connections, 3-18, 3-34, 3-36
scale reference menu, 6-38
error-correction for re ection
menu map, 8-12
searching for values with markers, 2-22
function block location, 1-4
response and isolation calibration, 6-73
second portion of conversion compression
restarting a calibration, 6-75
of memory data, 4-33

selecting line types, 4-14
shortened sweep time, 5-50
sensor calibration data, 5-34
signal ow in a mixer example, 3-33
menu map, 8-13
single bus concept, 11-19
single page of values printed or plotted, 4-30
single point limits, 2-41
six gigahertz operation option, 1-12
sloping limit lines, 2-39
sequence of measurement, 2-3
Smith chart markers, 2-15
cascading multiple sequences, 2-61
changing the sequence title, 2-58
clearing a sequence from memory, 2-57
decision making functions, 6-142
Smith or polar format markers, 2-12
Smo status notation, 1-8
editing a sequence, 2-56
gosub sequence command, 6-140
labels of display, 1-9
inserting a command, 2-56
limit test sequence, 2-64
softkey de nitions, 9-2
loading a sequence from disk, 2-60
loop counter sequence, 2-62
softkey menu maps, 8-1
softkeys and corresponding front panel access
modifying a command, 2-57
printing a sequence, 2-60
purging a sequence from disk, 2-60
reading using HP-IB, 6-144
running a test sequence, 2-55
for printing or plotting problems, 4-32
special functions menu, 6-142
stopping a sequencing, 2-55
disk storage problems, 4-43
storing a sequence on a disk, 2-59
TTL input for decision making, 6-140
uncorrected speci cations, 7-2
TTL I/O menu, 6-140
TTL out menu, 6-142
frequency:decrease, 5-50
with loop counter, 6-139
span setting for gating, 2-70
serial number plate location, 1-10
serial (RS-232) interface, 1-10
service and support options, 1-13
3.5 mm test ports, 7-8
service request asserted by the analyzer (S),
50 type-N test ports, 7-5
75 type-F test ports, 7-14
service request (SRQ) control line, 11-17
75 type-N test ports, 7-11
setting gate parameters, 2-70
7 mm test port, 7-2
setting HP-IB addresses, 11-20
setting measurement parameters using
external source mode, 7-29
sex of type-N calibration standard, 5-6
SH1 (full-source handshake), 11-18
frequency o set, 7-29
shape of gate, 2-70

harmonics (option 002), 7-28
* status notation, 1-8
steps of making a measurement, 2-3
spectral purity characteristics, 7-27
type-N test ports, 7-4
uncorrected test port, 7-2
setting with markers, 2-17
o set menus, 6-80
spectral purity characteristics, 7-27
a print or plot, 4-30
spectrum of RF ,LO and IF signals present in
a conversion loss measurement, 3-7
stop values possible, 1-7
SR1 (complete service request capabilities),
a sequence on a disk, 2-59
SRQ (service request) control line, 11-17
to external disk, 11-5
S (service request asserted by the analyzer),
to internal disk, 11-5
to internal memory, 11-5
full two port error-correction, 5-20
one port re ection error-correction, 5-18
structure of HP-IB bus, 11-16
subtract memory trace from the measurement
response and isolation error-correction
for& re ection measurements, 5-14
support and service options, 1-13
response and isolation error-correction
for& transmission measurements, 5-16
de nition of, 6-5
response error-correction for re ection
how to set auto sweep time, 5-51
how to set chop sweep, 5-54
response error-correction for transmission
de ning menus, 6-78
standard system maintenance service option,
start frequency set with markers, 2-17
how to set, 5-53
start values possible, 1-6

swept harmonics measurement, 2-66
swept IF conversion loss measurement
3.5 mm test port speci cations, 7-8
50 type-N test port speci cations, 7-5
swept power conversion compression
75 type-F test port speci cations, 7-14
75 type-N test port speci cations, 7-11
swept RF/IF mixer measurement, 3-7
test sequence connector location, 1-11
syntax for commands, 11-21
test sequence output (TEST SEQ), 7-35
cascading multiple sequences, 2-61
changing the sequence title, 2-58
clearing a sequence from memory, 2-57
decision making functions, 6-142
editing a sequence, 2-56
gosub sequence command, 6-140
how to change, 5-56
how to widen, 5-52
system controller capabilities (C1,C2,C3),
inserting a command, 2-56
limit test sequence, 2-64
system-controller mode, 11-19, 11-20
loading a sequence from disk, 2-60
loop counter sequence, 2-62
menu map, 8-15
modifying a command, 2-57
printing a sequence, 2-60
T6 (basic talker), 11-18
talker interface function, 11-15
purging a sequence from disk, 2-60
reading using HP-IB, 6-144
running a sequence, 2-55
special functions menu, 6-142
talk mode (T), 11-19
stopping a sequence, 2-55
target amplitude search, 2-24
storing a sequence on a disk, 2-59
TE0 (no extended talker capabilities), 11-18
TTL I/O menu, 6-140
TTL out menu, 6-142
with loop counter, 6-139
testing with limit lines, 2-36
test set switch, controlling the, 5-54
test using limits, 2-43
speci cations:50 , 7-2
speci cations:7 mm, 7-2
test port bias input (BIAS CONNECT), 7-35
test port coupling, 6-15
bandpass mode:re ection, 6-119
test port dimensions, 7-33
bandpass mode:transmission, 6-121

forward transform mode, 6-118
time domain bandpass, 6-117
time domain low pass impulse, 6-117
time domain low pass step, 6-117
transistor test xtures, 11-3
low pass impulse mode, 6-117
low pass mode , fault location
using bandpass mode, 6-121
using low pass mode, 6-125
low pass mode , re ection, 6-123
transmission response in time domain, 2-68
low pass mode , transmission, 6-125
low pass , setting frequency range, 6-122
uncorrected speci cations, 7-2
low pass step mode, 6-117
tri-state drivers (E2), 11-18
re ection response, 2-73
relative velocity factor, 6-119
fabricating and de ning standards, 6-91
how it works, 6-86
improving raw load match, 6-88
transmission response measurement, 2-68
improving raw source match, 6-88
time for measurements, 7-33
reason for using, 6-85
time stamp, 4-30, 11-13
requirements for standards, 6-90
area of display, 1-8
TRL*/LRM* two-port calibration, 6-74
TRL standard modi cation, 5-28
the active channel display, 2-8
TRM standard modi cation, 5-30
the displayed measurement, 4-29
to use fast 2-port calibration, 5-54
solving for printing or plotting, 4-32
with disk storage, 4-43
tsH status notation, 1-8
divide measurement data by the memory
T (talk mode), 11-19
trace math operation, 6-6
tuned receiver mode, 3-17, 6-111
compatible sweep types, 2-51
external source requirements, 2-52
transfer of data, 11-16
type-F (75 ) test ports
measurement uncertainties, 7-14, 7-17,
CW time-to-frequency domain, 6-134
type-N (50 ) test ports

type-N (75 ) test ports
type-N calibration standard sex, 5-6
type-N test ports (50 )
how to set, 5-53
typical calibration kit standard and
velocity factor for time domain
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
REF=FIXED MKR , 2-12
veri cation kit, 11-2
verifying performance of calibration, 6-84
3.5 mm test port measurements, 7-8
video output (EXT MON), 7-35
50 type-N test port measurements, 7-5
75 type-F test ports, 7-14, 7-17, 7-20,
both channels with a split display, 2-5
both measurement channels, 2-5
75 type-N test ports, 7-11
plot les on a PC, 4-20
7 mm test port measurements, 7-2
the measurement data and memory trace,
voltage selector switch, 1-10
" status notation, 1-8
up converter port connections, 3-3
widen system bandwidth, 5-52