Agilent Technologies
8719ET/20ET/22ET 8719ES/20ES/22ES Network Analyzers

© Copyright 1999–2002 Agilent Technologies, Inc.
Printed in USA

The information contained in this document is subject to change without notice.
Agilent Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Agilent Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
Agilent Technologies certifies that this product met its published specifications at the time of shipment from the factory. Agilent Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology, to the extent allowed by the Institute’s calibration facility, and to the calibration facilities of other International Standards Organization members.
The regulatory information is located in Chapter 8 , “Safety and Regulatory Information.”
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, AGILENT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. AGILENT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE OR PERFORMANCE OF THIS DOCUMENT OR ANY INFORMATION CONTAINED HEREIN. SHOULD AGILENT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL.
Product maintenance agreements and other customer assistance agreements are available for Agilent Technologies products. For any assistance, contact your nearest Agilent Technologies sales or service office. See Table 8-1 for the nearest office.

The following safety notes are used throughout this manual. Familiarize yourself with each of the notes and its meaning before operating this instrument. All pertinent safety notes for using this product are located in Chapter 8 , “Safety and Regulatory Information.”
This guide uses the following conventions:
This represents a key physically located on the instrument.
This represents a “softkey,” a key whose label is determined by the instrument’s firmware.
Screen Text This represents text displayed on the instrument’s screen.
Front-Panel Key

The Installation and Quick Start Guide provides procedures for installing, configuring, and verifying the operation of the analyzer. It also will help you familiarize yourself with the basic operation of the analyzer.
The User’s Guide shows how to make measurements, explains commonly-used features, and tells you how to get the most performance from your analyzer.
The Reference Guide provides reference information, such as specifications, menu maps, and key definitions.
The Programmer’s Guide provides general GPIB programming information, a command reference, and example programs. The Programmer’s Guide contains a CD-ROM with example programs.
The CD-ROM provides the Installation and Quick Start Guide, the User’s Guide, the Reference Guide, and the Programmer’s Guide in PDF format for viewing or printing from a PC.
The Service Guide provides information on calibrating, troubleshooting, and servicing your analyzer. The Service Guide is not part of a standard shipment and is available only as Option 0BW, or by ordering part number 08720-90397. A CD-ROM with the Service Guide in PDF format is included for viewing or printing from a PC.



Making Measurements Using This Chapter
This chapter contains the following example procedures for making measurements. Mixer and time domain measurements are covered in Chapter 2 , "Making Mixer Measurements (Option 089 Only)" and Chapter 3 , “Making Time Domain Measurements.” This chapter also describes how to use most display, marker, and sequencing functions.
— Phase Distortion (deviation from linear phase, group delay)
— Measuring Gain Compression
— Measuring Gain Compression and Reverse Isolation Simultaneously
— Making High Power Measurements (ES Analyzers Only)
The following chapters describe how to use more instrument functions (as indicated by their chapter titles):
— Electrical Length · (ES Analyzers Only)

Making Measurements More Instrument Functions Not Described in This Guide
To learn about instrument functions not covered in this user’s guide, refer to the following chapters in the reference guide.
“Menu Maps” contains maps of the instrument menu structure.
“Hardkey/Softkey Reference” contains descriptions of all instrument functions.

Making Measurements Making a Basic Measurement
There are five basic steps when you are making a measurement.
This example procedure shows you how to measure the transmission response of a bandpass filter.
Make the connections as shown in Figure 1-1.
Figure 1-1 Basic Measurement Setup

Press .
To set preset the analyzer to the “Factory Preset” conditions, press the
softkey if it is not selected. Then press .
Preset

Making Measurements Making a Basic Measurement
Setting the Frequency Range
To set the center frequency to 134 MHz, press:
Center 134 M/µ
To set the span to 30 MHz, press:

Setting the Source Power
To change the power level to −5 dBm, press:

To change the number of measurement data points to 101, press:
Sweep Setup NUMBER OF POINTS
To select the transmission measurement, press:
or on ET models:
To view the data trace, press:
Refer to the Chapter 5 , “Optimizing Measurement Results,” for procedures on correcting measurement errors.
To save the instrument state and error-correction in the analyzer internal memory, press:
Save/Recall SELECT DISK INTERNAL MEMORY RETURN SAVE STATE
Replace any standard used for error-correction with the device under test.
To measure the insertion loss of the bandpass filter, press:
Setting the Measurement · Scale Ref AUTOSCALE

Making Measurements Making a Basic Measurement
To create a printed copy of the measurement results, press:
Copy PRINT MONOCHROME PLOT
Refer to Chapter 4 , “Printing, Plotting, and Saving Measurement Results,” for procedures on how to set up a printer and define a print, plot, or save results.
(or )


This measurement example shows you how to measure the maximum amplitude of a surface acoustic wave (SAW) filter and then how to view the measurement data in the phase format, which provides information about the phase response.
Figure 1-2 Device Connections for Measuring a Magnitude Response


You may also want to select settings for the number of data points, averaging, and IF bandwidth.
Chan 1 Cal CALIBRATE MENU RESPONSE THRU
Press .
Preset

Making Measurements Measuring Magnitude and Insertion Phase Response
If the channels are coupled (the default condition), this calibration is valid for both channels.
Scale Ref AUTO SCALE
Figure 1-3 Example Magnitude Response Measurement Results

Display DUAL | QUAD SETUP DUAL CHAN ON
The channel 2 portion of Figure 1-4 shows the insertion phase response of the device under test. The analyzer measures and displays phase over the range of −180° to +180°. As phase changes beyond these values, a sharp 360° transition occurs in the displayed data.
Chan 2 · Format PHASE

Making Measurements Measuring Magnitude and Insertion Phase Response
Figure 1-4 Example Insertion Phase Response Measurement

The phase response shown in Figure 1-5 is undersampled; that is, there is more than 180° phase delay between frequency points. If the ∆Φ ≥ 180°, incorrect phase and delay information may result. Figure 1-5 shows an example of phase samples being with ∆Φ less than 180° and greater than 180°.
Figure 1-5 Phase Samples
Undersampling may arise when measuring devices with long electrical length. To correct this problem, the frequency span should be reduced, or the number of points increased until ∆Φ is less than 180° per point. Electrical delay may also be used to compensate for this effect (as shown in the next example procedure).

Making Measurements Using Display Functions
This section provides the necessary information for using the display functions. These functions are very helpful for displaying measurement data so that it will be easy to read. This section covers the following topics:

Making Measurements Using Display Functions
a. Turn the front panel knob to move the arrow pointer to the first character of the
title.
b. Press .
c. Repeat the previous two steps to enter the rest of the characters in your title. You
can enter a title that has a maximum of 50 characters.
d. Press to complete the title entry.
Figure 1-6 Example of a Display Title

Display MORE TITLE

Making Measurements Using Display Functions
In some cases, you may want to view more than one measured parameter at a time. Simultaneous gain and phase measurements, for example, are useful in evaluating stability in negative feedback amplifiers. You can easily make such measurements using the dual channel display.
, set to ON, and to 1X.
Display DUAL | QUAD SETUP DUAL CHAN on OFF SPLIT DISP
Figure 1-7 Example of Viewing Channel 1 and 2 Simultaneously


Making Measurements Using Display Functions
Figure 1-8 Example Dual Channel with Split Display On

Sweep Setup COUPLED CH OFF
Dual Channel Mode with Decoupled Stimulus
The stimulus functions of the two channels can be controlled independently using
in the stimulus menu. In addition, the markers can be controlled independently for each channel using in the marker mode menu, under the key.
COUPLED CH ON off
Dual Channel Mode with Decoupled Channel Power
By decoupling the channel power or port power and using the dual channel mode, you can simultaneously view two measurements (or two sets of measurements, if both auxiliary channels are enabled) having different power levels.
Marker Fctn

Making Measurements Using Display Functions
However, there are two configurations that will not sweep continuously.
1. For analyzers with source attenuators, with channel 1 having one attenuation value and channel 2 set to a different attenuation value, then continuous sweep is disabled to avoid wear on the attenuator. A similar situation where this occurs is when a 2-port cal is active and the port 1 attenuation value is not equal to the port 2 attenuation value. Since one attenuator is used for both measurements, this would cause the attenuator to continuously switch power ranges, so continuous sweep is not allowed. (The exception is analyzers configured with option 400. Option 400 analyzers have two attenuators and can allow different attenuation settings on each port.)
If one of these conditions exist, the test set hold mode will engage, and the status notation tsH will appear on the left side of the screen. The hold mode leaves the measurement function in only one of the two measurements. To update both measurement setups, press
Sweep Setup MEASURE RESTART
. Refer to "Source Attenuator Switch Protection" on page 7-13.
Four measurement channels can be viewed simultaneously by enabling auxiliary channels 3 and 4. Although independent of other channels in most variables, channels 3 and 4 are permanently coupled to channels 1 and 2 respectively by stimulus. That is, if channel 1 is set for a center frequency of 200 MHz and a span of 50 MHz, channel 3 will have the same stimulus values.
Channel 3 or 4 are activated when the Chan 3 or Chan 4 keys are pressed. Alternatively, you can enable the auxiliary setting to ON. For example, if channel 1 is active, pressing to ON enables channel 3 and its trace appears on the display. Channel 4 is similarly enabled and viewed when channel 2 is active.
Display DUAL | QUAD SETUP DUAL CHAN AUX CHAN SPLIT DISP 4X
The display will appear as shown in Figure 1-9. Channel 1 is in the upper-left quadrant of the display, channel 2 is in the upper-right quadrant, and channel 3 is in the lower half of the display.
Format · Chan 1

Making Measurements Using Display Functions
Figure 1-9 Three-Channel Display

Chan 2 AUX CHAN
This enables channel 4 and the screen now displays four separate grids as shown in Figure 1-10. Channel 4 is in the lower-right quadrant of the screen.

Making Measurements Using Display Functions
Figure 1-10 Four-Channel Display

Observe that the amber LED adjacent to the key is lit and the CH4 indicator on the display has a box around it. This indicates that channel 4 is now active and can be configured.
Marker MARKER 1 MARKER 2
Markers 1 and 2 appear on all four channel traces. Rotating the front panel control knob moves marker 2 on all four channel traces. Note that the active function, in this case the marker frequency, is the same color and in the same grid as the active channel (channel 4).
Observe that the amber LED adjacent to the key is lit. This indicates that channel 3 is now active and can be configured.
Press , set to UNCOUPLED.
Marker Fctn MARKER MODE MENU MARKERS:
Rotate the front panel control knob. Marker 2 moves only on the channel 3 trace.

Making Measurements Using Display Functions
Once made active, a channel can be configured independently of the other channels in most variables except stimulus. For example, once channel 3 is active, you can change its format to a Smith chart by pressing .
When one or both auxiliary channels are enabled, and interact to produce different display configurations according to Table 1-1.
DUAL CHAN on OFF SPLIT DISP 1X 2X 4X
Table 1-1 Customizing the Display
Split Display Dual Channel Aux Channels On Number of Graticules
1X Don’t Care Don’t Care 1
2X/4X Off 3 or 4 2
2X On Don’t Care
4X On 3 or 4 3
4X On Both on 4
gives you options for arranging the display of the channels. Press , to use .
Display DUAL|QUAD SETUP CHANNEL POSITION
works with . When is selected, gives you two choices for a two-graticule display:
CHANNEL POSITION SPLIT DISP 1X 2X 4X SPLIT DISP 2X
When is selected, gives you two choices for a four-graticule display:
Format SMITH CHART · 1X/2X/4X Off None · Channel Position Softkey

Making Measurements Using Display Functions
4 Param Displays Softkey
The menu does two things:
Figure 1-11 shows the first screen. Six setup options are described with softkeys through . is a four-parameter display where each channel is displayed on its own grid. Pressing immediately produces a four-grid, four-parameter display. is also a four-parameter display, except that channel 1 and channel 2 are overlaid on the upper grid and channel 3 and channel 4 are overlaid on the lower grid. The other setup softkeys operate similarly. Notice that setups D and F produce displays which include Smith charts.
4 PARAM DISPLAYS SETUP A SETUP F SETUP A
Pressing opens a screen which lists the order of keystrokes you would have to enter in order to create some of the setups without using one of the setup softkeys. The keystroke entries are listed (from top to bottom) beneath each setup and are color-coded to show the relationship between the keys and the channels. For example, beneath the four-grid display, [CHAN 1] and [MEAS] S11 are shown in yellow. Notice that in the four-grid graphic, Ch1 is also yellow, indicating that the keys in yellow apply to channel 1.
Pressing opens a screen which lists the hardkeys and softkeys associated with the auxiliary channels and setting up multiple-channel, multiple-grid displays. Next to each key is a description of its function.
Figure 1-11 4 Param Displays Menu

Display

Making Measurements Using Display Functions
The analyzer has four available memory traces, one per channel. Memory traces are totally channel dependent: channel 1 cannot access the channel 2 memory trace or vice versa. Memory traces can be saved with instrument states: one memory trace can be saved per channel for each saved instrument state. There are up to 31 save/recall registers available, so the total number of memory traces that can be present is 128 including the four active for the current instrument state. The memory data is stored as full precision, complex data. Memory traces must be displayed in order to be saved with instrument states.
Additional data can be stored onto 3.5-inch floppy disks using the front panel disk drive.
Two trace math operations are implemented:
(Note that normalization is not .) Memory traces are saved and recalled and trace math is done immediately after error-correction. This means that any data processing done after error-correction, including parameter conversion, time domain transformation (Option 010), scaling, etc., can be performed on the memory trace. You can also use trace math as a simple means of error-correction, although that is not its main purpose.
All data processing operations that occur after trace math, except smoothing and gating, are identical for the data trace and the memory trace. If smoothing or gating is on when a memory trace is saved, this state is maintained regardless of the data trace smoothing or gating status. If a memory trace is saved with gating or smoothing on, these features can be turned on or off in the memory-only display mode.
The actual memory for storing a memory trace is allocated only as needed. The memory trace is cleared on instrument preset, power on, or instrument state recall.
If sweep mode or sweep range is different between the data and memory traces, trace math is allowed, and no warning message is displayed. If the number of points in the two traces is different, the memory trace is not displayed nor rescaled. However, if the number of points for the data trace is changed back to the number of points in the memory, the memory trace can then be displayed.
If trace math or display memory is requested and no memory trace exists, the message CAUTION: NO VALID MEMORY TRACE is displayed.
To Save a Data Trace to the Display Memory
Press to store the current active measurement data in the memory of the active channel. The data trace is now also the memory trace. You can use a memory trace for subsequent math manipulations.
Display DATA→MEMORY

Making Measurements Using Display Functions
To View the Measurement Data and Memory Trace
The analyzer default setting shows you the current measurement data for the active channel.
This is the only memory display mode where you can change the smoothing and gating of the memory trace.
Display DATA and MEMORY
To Divide Measurement Data by the Memory Trace
You can use this feature for ratio comparison of two traces, for example, measurements of gain or attenuation.
The analyzer normalizes the data to the memory, and shows the results.
To Subtract the Memory Trace from the Measurement Data Trace
You can use this feature for storing a measured vector error, for example, directivity. Then, you can later subtract it from the device measurement.
The analyzer performs a vector subtraction on the complex data.
To Ratio Measurements in Channel 1 and 2
You may want to use this feature when making amplifier measurements to produce a trace that represents gain compression. For example, with the channels uncoupled, you can increase the power for channel 2 while channel 1 remains unchanged. This will allow you to observe the gain compression on channel 2.
Sweep Setup COUPLED CH OFF
Display MORE D2/D1 TO D2 ON
Display MEMORY · Display DATA/MEM · Display DATA-MEM

Making Measurements Using Display Functions
Pressing switches off the analyzer display while leaving the instrument in its current measurement state. This feature may be helpful in prolonging the life of the LCD in applications where the analyzer is left unattended (such as in an automated test system). Turning the front panel knob or pressing any front panel key will restore normal display operation.
Display ADJUST DISPLAY BLANK DISPLAY
Pressing will blank the displayed frequency notation for security purposes. The frequency labels cannot be restored except by instrument preset or turning the power off and then on.
Display FREQUENCY BLANK

Making Measurements Using Display Functions
To adjust the intensity of the display, press and rotate the front panel knob, use the ( ) ( ) keys, or use the numerical keypad to set the intensity value between 50 and 100 percent. Lowering the intensity may prolong the life of the LCD.
Display ADJUST DISPLAY INTENSITY
To set all the display elements to the factory-defined default colors, press
The Modify Colors Menu
The softkey within the adjust display menu provides access to the modify colors menu.
The modify colors menu allows you to adjust the colors on your analyzer’s display. The default colors in this instrument were chosen to maximize your ability to discern the difference between the channel colors, and to comfortably and effectively view the colors. Each channel’s memory trace color was chosen because the color is similar to the channels data trace color. This allows easy association between the data trace and the memory trace for each channel.
You may choose to change the default colors to suit environmental needs, individual preferences, or to accommodate color deficient vision. You can use any of the available colors for any of the display elements listed:
CH1 MEM CH3 MEM
CH2 DATA/LIMIT LN CH4 DATA/LIMIT LN
CH2 MEM CH4 MEM
GRATICULE REF LINE
TEXT WARNING
To change the color of a display elements, press the softkey for that element (such as
). Then press and turn the analyzer front panel knob; use the step keys or the numeric keypad, until the desired color appears.
Setting Display Intensity · Setting Default Colors · Preset

Making Measurements Using Display Functions
Table 1-2 Display Colors with Maximum Viewing Angle
Display Color Tint Brightness Color
Red 0 100 100
Yellow 17 100 100
Green 33 100 100
Cyan 50 100 100
Blue 67 100 100
Magenta 83 100 100
White N/A 100 0
Color is comprised of three parameters:
The most frequently occurring color deficiency is the inability to distinguish red, yellow, and green from one another. Confusion between these colors can usually be eliminated by increasing the brightness between the colors. To accomplish this, press the
softkey and turn the analyzer front panel knob. If additional adjustment is needed, vary the degree of whiteness of the color. To accomplish this, press the softkey and turn the analyzer front panel knob.
COLOR
To save a modified color set, press . Modified colors are not part of a saved instrument state and are lost unless saved using these softkeys. Once modified colors are saved, they will be the colors applied until is pushed.
To recall the previously saved color set, press .

Making Measurements Using Markers
The key displays a movable active marker on the screen and provides access to a series of menus to control up to five display markers for each channel. Markers are used to obtain numerical readings of measured values. They also provide capabilities for reducing measurement time by changing stimulus parameters, searching the trace for specific values, or statistically analyzing part or all of the trace.
Markers have a stimulus value (the x-axis value in a Cartesian format) and a response value (the y-axis value in a Cartesian format). In polar format, the second part of a complex data pair is also provided as an auxiliary response value. In Smith chart format, the real and imaginary rectangle are both displayed, and the effective capacitance or inductance of the imaginary part is also displayed. When a marker is activated and no other function is active, its stimulus value is displayed in the active entry area and can be controlled with the knob, the step keys, or the numeric keypad. The active marker can be moved to any point on the trace, and its response and stimulus values are displayed at the top right corner of the graticule for each displayed channel, in units appropriate to the display format. The displayed marker response values are valid even when the measured data is above or below the range displayed on the graticule.
Marker values are normally continuous: that is, they are interpolated between measured points. They can also be set to read only discrete measured points. Markers normally have the same stimulus values for all channels, or they can be uncoupled so that each channel has independent markers, regardless of whether stimulus values are coupled or dual channel display is on.
The analyzer can either place markers on discrete measured points, or move the markers continuously along a trace by interpolating the data value between measured points.
Marker Fctn MARKER MODE MENU
— Choose if you want the analyzer to place markers at any point on the trace, by interpolating between measured points. This default mode allows you to conveniently obtain round numbers for the stimulus value.
— Choose if you want the analyzer to place markers only on measured trace points determined by the stimulus settings. This may be the best mode to use with automated testing, using a computer or test sequencing because the analyzer does not interpolate between measured points.
Marker

To Activate Display Markers
The active marker is identified on the analyzer display with the following symbol: ∇
The active marker stimulus value is displayed in the active entry area. You can modify the stimulus value of the active marker, using the front panel knob or numerical keypad. All of the marker response and stimulus values are displayed in the upper right corner of the display.
Figure 1-12 Active Marker Control Example

To switch on the corresponding marker and make it the active marker, press:
All of the markers, other than the active marker, become inactive and are represented on the analyzer display as ∆. The active and inactive markers are shown in Figure 1-13.

Making Measurements Using Markers
Figure 1-13 Active and Inactive Markers Example

To Move Marker Information Off the Grids
If marker information obscures the display traces, you can turn off the softkey menu and move the marker information off the display traces and into the softkey menu area. Pressing the backspace key performs this function. This is a toggle function. Pressing alternately hides and restores the current softkey menu. The softkey menu is also restored when you press any softkey or a hardkey which leads to a menu.
Marker 1 2 3 4
Marker Fctn MARKER MODE MENU MARKERS:
The display will be similar to Figure 1-14.
Marker

Figure 1-14 Marker Information Moved into the Softkey Menu Area

The display will be similar to Figure 1-15.
Making Measurements · Using Markers · pg654e

Making Measurements Using Markers
Figure 1-15 Marker Information on the Graticules

You can also restore the softkey menu by pressing a hardkey which opens a menu (such as
) or pressing a softkey.
To Use Delta (∆) Markers
This is a relative mode, where the marker values show the position of the active marker relative to the delta reference marker. You can switch on the delta mode by defining one of the five markers as the delta reference.
Marker ∆ MODE MENU ∆ REF=1
pg655e · Meas

Figure 1-16 Marker 1 as the Reference Marker Example

To Activate a Fixed Marker
When a reference marker is fixed, it does not rely on a current trace to maintain its fixed position. This is convenient when comparing two different measurement conditions. To activate a fixed marker on the analyzer, press .
Marker zero puts a fixed reference at the current position of the active marker.
Marker ∆ MODE MENU
To change to a Delta Marker to a fixed reference marker, press .
Using the Key to Activate a Fixed Reference Marker
Marker zero enters the position of the active marker as the ∆ reference position. Alternatively, you can specify the fixed point with . Marker zero is canceled by switching delta mode off.
and turn the front panel knob, or enter a value from the front panel keypad.
and turn the front panel knob, or enter a value from the front panel keypad.
and turn the front panel knob, or enter a value from the front panel keypad.
∆MODE MENU FIXED MKR POSITION FIXED MKR STIMULUS

Making Measurements Using Markers
Figure 1-17 Example of a Fixed Reference Marker Using MKR ZERO

Using the Key to Activate a Fixed Reference Marker
and turn the front panel knob, or enter a value from the front panel keypad.
Marker ∆MODE MENU ∆REF=∆FIXED MKR ∆MODE MENU FIXED MKR POSITION FIXED MKR STIMULUS
The marker is shown on the display as a small delta (∆), smaller than∆ the inactive marker triangles.
and turn the front panel knob, or enter a value from the front panel keypad.
In a Cartesian format, the setting is the y-axis value. In polar or Smith chart format, with a magnitude/phase marker, a real/imaginary marker, an R+jX marker, or a G+jB marker, the setting applies to the first part of the complex data pair. (Fixed marker response values are always uncoupled in the two channels.)
and turn the front panel knob, or enter a value from the front panel keypad.
This value is the second part of complex data pair, and applies to a magnitude/phase marker, a real/imaginary marker, an R+jX marker, or a G+jB marker. (Fixed marker auxiliary response values are always uncoupled in the two channels.)

Figure 1-18 Example of a Fixed Reference Marker Using (∆)REF=(∆)FIXED MKR

At a preset state, the markers have the same stimulus values on each channel, but they can be uncoupled so that each channel has independent markers.
Press and select from the following keys:
Marker Fctn MARKER MODE MENU
Figure 1-19 Example of Coupled and Uncoupled Markers

Making Measurements · Using Markers

Making Measurements Using Markers
The analyzer can display the marker value as magnitude and phase, or as a real/imaginary pair: gives linear magnitude and phase, gives log magnitude and phase, gives the real value first, then the imaginary value.
LIN MKR LOG MKR Re/Im
You can use these markers only when you are viewing a polar display format. (The format is available from the key.)
Marker Fctn MKR MODE MENU MARKERS:DISCRETE
Format POLAR Marker Fctn MARKER MODE MENU POLAR MKR MENU
Figure 1-20 Example of a Log Marker in Polar Format

Format · Re/Im MKR

For greater accuracy when using markers in the Smith chart format, activate the discrete marker mode. Press .
Marker Fctn MKR MODE MENU MARKERS:DISCRETE
To use Smith chart format:
Marker Fctn MARKER MODE MENU SMITH MKR MENU
The marker annotation tells that the complex impedance is capacitive in the bottom half of the Smith chart display and is inductive in the top half of the display.

Making Measurements Using Markers
Figure 1-21 Example of Impedance Smith Chart Markers

The analyzer allows you to set measurement parameters with the markers, without going through the usual key sequence. You can change certain stimulus and response parameters to make them equal to the current active marker value.
Setting the Start Frequency
Figure 1-22 Example of Setting the Start Frequency Using a Marker

Marker Fctn

Setting the Stop Frequency
Figure 1-23 Example of Setting the Stop Frequency Using a Marker

Setting the Center Frequency

Making Measurements Using Markers
Figure 1-24 Example of Setting the Center Frequency Using a Marker

Setting the Frequency Span
You can set the span equal to the spacing between two markers. If you set the center frequency before you set the frequency span, you will have a better view of the area of interest.
Marker ∆MODE MENU ∆REF=1 MARKER 2
Iterate between marker 1 and marker 2 by pressing and , respectively, and turning the front panel knob or entering values from the front panel keypad to position the markers around the center frequency. When finished positioning the markers, make sure that marker 2 is selected as the active marker.
Marker Fctn MARKER→SPAN

Figure 1-25 Example of Setting the Frequency Span Using Marker

Setting the Display Reference Value
Figure 1-26 Example of Setting the Reference Value Using a Marker

Setting the Electrical Delay
This feature adds phase delay to a variation in phase versus frequency, therefore it is only applicable for ratioed inputs.
Making Measurements · Using Markers · Marker Fctn

Making Measurements Using Markers
Additional electrical delay adjustments are required on devices without constant group delay over the measured frequency span.
Figure 1-27 Example of Setting the Electrical Delay Using a Marker

and either turn the front panel knob or enter the value, followed by a unit terminator.
Seq SPECIAL FUNCTIONS MKR→CW
You can use this function to set the marker to a gain peak in an amplifier. After pressing
, activate a CW frequency power sweep to look at the gain compression with increasing input power.
Format PHASE · Marker Fctn · Marker

These functions place the marker at an amplitude-related point on the trace. If you switch on tracking, the analyzer searches every new trace for the target point.
Searching for the Maximum Amplitude
Figure 1-28 Example of Searching for the Maximum Amplitude Using a

Searching for the Minimum Amplitude

Making Measurements Using Markers
Figure 1-29 Example of Searching for the Minimum Amplitude Using a Marker

Searching for a Target Amplitude
Marker Search TARGET VALUE
to enter the new value.
Figure 1-30 Example of Searching for a Target Amplitude Using a Marker

Marker Search

Searching for a Bandwidth
The analyzer can automatically calculate and display the bandwidth (BW:), center frequency (CENT:), Q, and loss of the device under test at the center frequency. (Q stands for “quality factor,” defined as the ratio of a circuit's resonant frequency to its bandwidth.) These values are shown in the marker data readout.
Figure 1-31 Example of Searching for a Bandwidth Using Markers

Tracking the Amplitude that You Are Searching
Marker Search TRACKING ON
When tracking is not activated, the analyzer finds the specified amplitude on the current sweep and the marker remains at same stimulus value, regardless of changes in the trace response value with subsequent sweeps.
Making Measurements · Using Markers · Marker Search

Making Measurements Using Markers
This function calculates the mean, standard deviation, and peak-to-peak values of the section of the displayed trace between the active marker and the delta reference. If there is no delta reference, the analyzer calculates the statistics for the entire trace.
Marker ∆ MODE MENU ∆ REF=1
Marker Fctn MKR MODE MENU MKR STATS ON
An application for this feature is to find the peak-to-peak value of passband ripple without searching separately for the maximum and minimum values.
If you are viewing a measurement in the polar or Smith Chart format, the analyzer calculates the statistics using the first value of the complex pair (magnitude, real part, resistance, or conductance).
Figure 1-32 Example Statistics of Measurement Data


Making Measurements Measuring Electrical Length and Phase Distortion
The analyzer mathematically implements a function similar to the mechanical “line stretchers” of earlier analyzers. This feature simulates a variable length lossless transmission line, which you can add to or remove from the analyzer's receiver input to compensate for interconnecting cables, etc. In this example, the electronic line stretcher measures the electrical length of a SAW filter.
The analyzer allows you to measure the linearity of the phase shift through a device over a range of frequencies and the analyzer can express it in two different ways:
Figure 1-33 Device Connections for Measuring Electrical Length
Meas Trans:FWD S21 (B/R) TRANSMISSN
or on ET models:
Center 134 M/µ
Span 2 M/µ
Scale Ref AUTO SCALE

Making Measurements Measuring Electrical Length and Phase Distortion
You may also want to select settings for the number of data points, averaging, and IF bandwidth.
Cal CALIBRATE MENU RESPONSE THRU
Scale Ref AUTO SCALE
The linearly changing phase is due to the device’s electrical length. You can measure this changing phase by adding electrical length (electrical delay) to compensate for it.
Figure 1-34 Linearly Changing Phase

Marker Fctn MARKER→DELAY
This function calculates and adds in the appropriate electrical delay by taking a ±10% span about the marker, measuring the ∆Φ, and computing the delay as the negative of ∆Φ / ∆ frequency.
Alternatively, press and turn the front panel knob to increase the electrical length until you achieve the best flat line, as shown in Figure 1-35.
Scale Ref ELECTRICAL DELAY
Marker

Making Measurements Measuring Electrical Length and Phase Distortion
The measurement value that the analyzer displays represents the electrical length of your device relative to the speed of light in free space. The physical length of your device is related to this value by the propagation velocity of its medium.
You could change the velocity factor to compensate for propagation velocity by pressing (enter the value) . This will allow the analyzer to accurately display the equivalent distance that corresponds to the entered electrical delay.
Cal MORE VELOCITY FACTOR x1
Figure 1-35 Example Best Flat Line with Added Electrical Delay

Scale Ref ELECTRICAL DELAY
In this example, there is a large amount of electrical delay due to the long electrical length of the SAW filter under test.
This portion of the example shows you how to measure the linearity of the phase shift over a range of frequencies. The analyzer allows you to measure this linearity and read it in two different ways: deviation from linear phase, or group delay.

Making Measurements Measuring Electrical Length and Phase Distortion
Deviation From Linear Phase
By adding electrical length to “flatten out” the phase response, you have removed the linear phase shift through your device. The deviation from linear phase shift through your device is all that remains.
Scale Ref SCALE DIV
Marker Fctn MKR MODE MENU STATS ON
Figure 1-36 Deviation From Linear Phase Example Measurement

The phase linearity of many devices is specified in terms of group or envelope delay. The analyzer can translate this information into a related parameter, group delay. Group delay is the transmission time through your device under test as a function of frequency. Mathematically, it is the derivative of the phase response which can be approximated by the following ratio:
where ∆Φ is the difference in phase at two frequencies separated by ∆F. The quantity ∆F is commonly called the “aperture” of the measurement. The analyzer calculates group delay from its phase response measurements.
Group Delay

Making Measurements Measuring Electrical Length and Phase Distortion
The default aperture is the total frequency span divided by the number of points across the display (i.e. 201 points or 0.5% of the total span in this example).
Format DELAY Scale Ref SCALE DIV
Figure 1-37 Group Delay Example Measurement

Group delay measurements may require a specific aperture (∆)F) or frequency spacing between measurement points. The phase shift between two adjacent frequency points must be less than 180°, otherwise incorrect group delay information may result.
When you increase the aperture, the analyzer removes fine grain variations from the response. It is critical that you specify the group delay aperture when you compare group delay measurements.
Marker · Avg SMOOTHING ON

Making Measurements Measuring Electrical Length and Phase Distortion
Figure 1-38 Group Delay Example Measurement with Smoothing

SMOOTHING APERTURE 5 x1
As the aperture is increased the “smoothness” of the trace improves markedly, but at the expense of measurement detail.
Figure 1-39 Group Delay Example Measurement with Smoothing Aperture

Group delay is calculated by dividing the phase difference between points by the frequency spacing. Thus, if n equals the number of points, the number of phase difference values (or frequency segments) will be n−1. The first data point is repeated so that the total number of points remains n.
Increased

Making Measurements Characterizing a Duplexer (ES Analyzers Only)
This measurement example demonstrates how to characterize a 3-port device, in this case a duplexer, using four-parameter display mode. You must use a test adapter or a special 3-port test adapter to route the signals from the analyzer (a two-port instrument) to the duplexer (a three-port device). This example procedure is performed using one of the following test adapters:
J 8720D Option K36 Duplexer Switching Test Set
J 8720D Option K39 3-Port Switching Test Set
The following abbreviations are used in reference to a duplexer:
Figure 1-40 Duplexer Connections


Making Measurements Characterizing a Duplexer (ES Analyzers Only)
Sweep Setup COUPLED CH on OFF OFF
System CONFIGURE MENU USER SETTINGS
K36 MODE on OFF Meas SELECT [TX-ANT]
K39 MODE on OFF Meas SELECT PORTS [1-3]
10.Set up channel 2 for the Ant-Rx stimulus parameters. In this example, a wide frequency range that covers both the Tx-Ant and Ant-Rx parameters has been chosen.
11.Set up control of the test adapter so that channels 2 and 4 are Rx:
Meas SELECT PORTS [2-3]
12.Perform a full two-port calibration on channel 2.
13.Save this state in the analyzer: Press .
14.Connect the duplexer to the test adapter.

Making Measurements Characterizing a Duplexer (ES Analyzers Only)
15.Set up a 2-graticule, 4-parameter display with transmission measurements on the top graticule and reflection measurements on the bottom graticule: Press
Display DUAL|QUAD SETUP 4-PARAM DISPLAYS SETUP B Meas Trans: REV S12 (A/R)

Chan 4 Refl: REV S22 (B/R) Chan 1 Trans:FWD S21 (B/R) Chan 3 Refl: FWD S11 (A/R) DUAL CHAN on OFF ON
The display will be similar to Figure 1-41.
Figure 1-41 Duplexer Measurement

Normally, a 2-port calibration requires a forward and reverse sweep to complete before the displayed trace updates. For faster tuning, it is possible to set the number of sweeps for the active display channel (S11 and S21 for channel 1 in this case) to update more often than the inactive display channel. In this example we choose 8 updates of the forward parameters to 1 update of the reverse in channel 1, and 8 updates of the reverse to 1 update of the forward in channel 2 (where the active parameters are S22 and S12).
Press . Press .
Chan 1 System CONFIGURE MENU TESTSET SW CONTINUOUS 8 x1
Chan 2 System CONFIGURE MENU TESTSET SW CONTINUOUS 8 x1

Making Measurements Measuring Amplifiers
The analyzer allows you to measure the transmission and reflection characteristics of many amplifiers and active devices. You can measure scalar parameters such as gain, gain flatness, gain compression, reverse isolation, return loss (SWR), and gain drift versus time. Additionally, you can measure vector parameters such as deviation from linear phase, group delay, complex impedance and AM-to-PM conversion. You can also make high power measurements.
Figure 1-42 Amplifier Parameters
When you are measuring a device that is very sensitive to absolute power level, it is important that you accurately set the power level at either the device input or output. The analyzer is capable of using an external GPIB power meter and controlling source power directly. Refer to Chapter 6 , "Calibrating for Increased Measurement Accuracy" for information on power meter calibration.
This section contains the following measurement examples:
— Option 085 Configuration
— Option 012 Configuration

Making Measurements Measuring Amplifiers
Gain compression occurs when the input power of an amplifier is increased to a level that reduces the gain of the amplifier and causes a nonlinear increase in output power. The point at which the gain is reduced by 1 dB is called the 1 dB compression point. The gain compression will vary with frequency, so it is necessary to find the worst-case point of gain compression in the frequency band.
Once that point is identified, you can perform a power sweep of that CW frequency to measure the input power at which the 1 dB compression occurs and the absolute power out (in dBm) at compression. The following steps provide detailed instruction on how to apply various features of the analyzer to accomplish these measurements.
NOTE In a compression measurement it is necessary to know the RF input or output power at a certain level of gain compression. Therefore, both gain and absolute power level need to be accurately characterized. Uncertainty in a gain compression measurement is typically less than 0.05 dB. Also, each input channel of the analyzer is calibrated to display absolute power (typically within +0.5 dBm up to 3 GHz, and +1 dB up to 6 GHz). This can be improved by calibrating the power meter. Refer to "Power Meter Measurement Calibration" on page 6-33 for information on calibrating the power meter.
Figure 1-43 Diagram of Gain Compression
Avg IF BW 1000 x1
Chan 1 Meas Trans:FWD S21 (B/R) TRANSMISSN
or on ET models:

Making Measurements Measuring Amplifiers
D1/D2 to D2 ON
Display DATA →MEMORY DATA/MEM
Display DUAL | QUAD SETUP DUAL CHANNEL ON
Chan 2 Meas Trans:FWD S21 (B/R) TRANSMISSN
Sweep Setup COUPLED CH OFF
This will allow you to separately increase the power for channel 2 and channel 1, so that you can observe the gain compression on channel 2 while channel 1 remains unchanged.
Chan 2 DISPLAY MORE D2/D1 to D2 ON
This produces a trace that represents gain compression only.
Scale Ref SCALE/DIV x1
10.Increase the power until you observe approximately 1 dB of compression on channel 2, using the step keys or the front panel knob.
11.To locate the worst case point on the trace, press:
Marker Search SEARCH:MIN
Marker MARKER 1 · Power


12.If was selected, recouple the channel stimulus by pressing:
Sweep Setup COUPLED CH ON
13.To place the marker exactly on a measurement point, press:
Marker Fctn MARKER MODE MENU MARKERS:DISCRETE
14.To set the CW frequency before going into the power sweep mode, press:
Seq SPECIAL FUNCTIONS MARKER→ CW
15.Press .
Sweep Setup SWEEP TYPE MENU POWER SWEEP
If interpolation is on (the default setting), the calibration will be applied to the power sweep.
16.Enter the start and stop power levels for the sweep.
Now channel 1 is displaying a gain compression curve. (Do not pay attention to channel 2 at this time.)

Chan 2 Display DUAL | QUAD SETUP DUAL CHANNEL ON
D2/D1 to D2 ON MORE D2/D1 to D2 OFF
Now channel 2 displays absolute output power (in dBm) as a function of power input.
20.Press to change the scale of channel 2 to 10 dB per division.
Scale Ref SCALE/DIV 10 x1
21.Press to change the scale of channel 1 to 1 dB per division.
Chan 1 1 x1

Making Measurements Measuring Amplifiers
Marker Fctn MARKER MODE MENU MARKERS:COUPLED
22.Press .
23.To find the 1 dB compression point on channel 1, press:
Marker Search SEARCH:MAX Marker MKR ZERO Marker Search
SEARCH:TARGET −1 x1
Chan 2 Marker MKR MODE MENU MARKERS:UNCOUPLED
24.Press .
25.To take the channel 2 marker out of the ∆ mode so that it reads the absolute output power of the amplifier (in dBm), press:
Marker ∆ MODE MENU ∆ MODE OFF
Figure 1-45 Gain Compression Using Power Sweep


Making Measurements Measuring Amplifiers
Since an amplifier will have high gain in the forward direction and high isolation in the reverse direction, the gain (S21) will be much greater than the reverse isolation (S12). Therefore, the power you apply to the input of the amplifier for the forward measurement (S21) should be considerably lower than the power you apply to the output for the reverse measurement (S12). By applying low power in the forward direction, you’ll prevent the amplifier from being saturated. A higher power in the reverse direction keeps noise from being a factor in the measurement and accounts for any losses caused by attenuators or couplers on the amplifier’s output needed to lower the output power into the analyzer. The following steps demonstrate the features that best accomplish these measurements.
Sweep Setup COUPLED CH ON
Coupling the channels allows you to have the same frequency range and calibration applied to channel 1 and channel 2.
Power PORT POWER [UNCOUPLED]
Uncoupling the port power allows you to apply different power levels at each port. In Figure 1-46, the port 1 power is set to −25 dBm for the gain measurement (S21) and the port 2 power is set to 0 dBm for the reverse isolation measurement (S12).
Chan 1 Meas Trans:FWD S21 (B/R) Power
Display DUAL | QUAD SETUP DUAL CHAN ON
You can view both measurements simultaneously by using the dual channel display mode. Refer to Figure 1-46. If the port power levels are in different power ranges, one of the displayed measurements will not be continually updated and the annotation tsH will appear on the left side of the display. Refer to "Source Attenuator Switch Protection" on page 7-13 for information on how to override this state.

Making Measurements Measuring Amplifiers
Figure 1-46 Gain and Reverse Isolation


Making Measurements Measuring Amplifiers
Analyzers equipped with Option 085 can be configured to measure high power devices. This ability is useful if the required input power for a device under test is greater than the analyzer can provide, or if the maximum output power from an amplifier under test exceeds safe input limits for a standard analyzer. This section describes how to set up the analyzer to perform high power measurements.
Figure 1-47 High Power Test Setup (Step 1)
Before continuing, save this state and rename it to UPRESET. To allow the preset mode to be set to user preset, press . The final state can then be resaved as a user preset to avoid an over-power condition from the factory preset.
PWR RANGE AUTO POWER RANGES −20 x1
Initial Setup · Determining Power Levels · Power

Making Measurements Measuring Amplifiers
10.At the maximum estimated power level, determine if the maximum output power from the coupled arm of the coupler will be higher than the acceptable limit. If so, add the appropriate amount of attenuation that will keep the coupled output power below −10 dBm and above −35 dBm.
11.Switch off the booster amplifier.
12.Make a connection between the open port of the 20 dB coupler and the RF IN connector on the rear panel of the analyzer.
13.Make a connection between the coupled arm of the 20 dB coupler (along with any added attenuation) and the R CHANNEL IN connector on the front panel.
Additional Setup

Making Measurements Measuring Amplifiers
Figure 1-48 High Power Test Setup (Step 2a)
Figure 1-49 High Power Test Setup (Step 2b)

Selecting Power Ranges and Attenuator Settings
14.Select a power range that will not exceed the maximum estimated power level that will force the DUT into compression. For example, if your booster amplifier has a gain of +15 dB and the DUT will compress if supplied with +20 dBm, then you would adjust the analyzer output power to not exceed +5 dBm by pressing ( , 8722ES).
Power PWR RANGE MAN POWER RANGES RANGE 0 −15 TO +5 RANGE 0 −20 TO −5
15.Adjust the analyzer power level to be below the maximum limit of the power range by pressing (or less) .

Making Measurements Measuring Amplifiers
16.Estimate the maximum amount of gain that could be provided by the DUT and, as a result, the maximum amount of power that could be received by TEST PORT 2 when the DUT is in compression. For example, if a DUT with a maximum gain of +10 dB receives an input of +20 dBm, then the maximum amount of power that could be received by TEST PORT 2 is +30 dBm.
Figure 1-50 Internal Signal Paths of the Option 085 Analyzer

17.Calculate the amount of attenuation needed between the analyzer’s couplers and samplers in order to not exceed the optimum sampler power level of −10 dBm.
In this example, it will be necessary to take the following into consideration:

Making Measurements Measuring Amplifiers
With the previous points in mind, the amount of attenuation can be calculated from the following equations:
18.Set the internal step attenuators to the values calculated in the previous step (rounding off to the highest 5 dB step). Press .
Power ATTENUATOR A 15 x1
ATTENUATOR B 25 x1
19.Switch on the booster amplifier.
20.To activate the external reference mode, press .
21.Measure the output power from test port 1 using a power meter and verify that it is as expected.
22.If you are measuring a highly reflective device, high power isolators should be inserted in place of the jumpers located between the two sets of front panel SWITCH and COUPLER connectors.
23.Confirm that all power and attenuator settings are correct, and set the following measurement parameters:
24.Perform a response calibration:
Cal CALIBRATE MENU RESPONSE THRU
25.Make the connections as shown in Figure 1-51. Switch on the DUT and measure the S21 gain of the amplifier under test to confirm the proper operation of the measurement test setup.

Making Measurements Measuring Amplifiers
Figure 1-51 High Power Test Setup (Step 3)

26.Make any other desired high power measurements.
Ratio measurements such as gain will be correctly displayed. However, the displayed absolute power levels on the analyzer will not be correct. To correctly interpret power levels, the gain of the booster amplifier and the attenuator settings must be taken into consideration.

Making Measurements Measuring Amplifiers
Analyzers equipped with Option 012 can be configured to measure devices that have high power outputs. With direct sampler access, you can insert attenuators between the couplers and samplers, protecting the samplers from excessive power.
Figure 1-52 Internal Signal Paths of the Option 012 Analyzer

In this example, it will be necessary to take the following into consideration:

Making Measurements Measuring Amplifiers
With the previous points in mind, the amount of attenuation can be calculated from the following equations:
Figure 1-53 Option 012 High Power Configuration


Making Measurements Using the Swept List Mode to Test a Device
When using a list frequency sweep, the analyzer has the ability to sweep arbitrary frequency segments, each containing a list of frequency points. One major advantage of using list frequency sweep is that it allows you to measure the minimum number of data points, and only at the frequencies of interest. This serves to minimize the overall test time. Two different list frequency sweep modes can be selected:
Stepped List Mode In this mode, the source steps to each defined frequency point, stopping while data is taken. This mode eliminates IF delay and allows frequency segments to overlap. However, the sweep time is substantially slower than for a continuous sweep with the same number of points.
Swept List Mode This mode takes data while sweeping through the defined frequency segments, increasing throughput by up to 6 times over a stepped sweep. In addition, this mode allows the test port power and IF bandwidth to be set independently for each segment that is defined. The frequency segments in this mode cannot overlap.
The ability to completely customize the frequency sweep while using swept list mode is useful when setting up a measurement for a device with high dynamic range, like a filter. The following measurement of a filter illustrates the advantages of using the swept list mode.
Figure 1-54 Swept List Measurement Setup

Making Measurements Using the Swept List Mode to Test a Device
or on ET models:
Center 900 M/µ
Span 500 M/µ
Figure 1-55 Characteristics of a Filter

Sweep Setup SWEEP TYPE MENU EDIT LIST

Making Measurements Using the Swept List Mode to Test a Device
Set Up the Lower Stopband Parameters

LIST POWER ON off SEGMENT POWER 10 x1
until ON is selected
LIST IF BW ON off SEGMENT IF BW 1000 x1
until ON is selected
Set Up the Passband Parameters


Set Up the Upper Stopband Parameters

![Page 84 of the Agilent Technologies Network Analyzer (20ES) user manual — DONE LIST FREQ [SWEPT]](/cdn/reader/15/150a289e983abebdc4bd6f0c3204b03f577bfd829d08c0cf5bb1353a14e103d7/images/page-84.webp?w=400)
Making Measurements Using the Swept List Mode to Test a Device
SEGMENT POWER 10 x1
SEGMENT IF BW 300 x1
Calibrate and Measure
Figure 1-56 Calibrated Swept List Thru Measurement
![Agilent Technologies Network Analyzer (20ES) — DONE LIST FREQ [SWEPT] diagram (page 84)](/cdn/reader/15/150a289e983abebdc4bd6f0c3204b03f577bfd829d08c0cf5bb1353a14e103d7/images/rimg-47.webp)
In Figure 1-57, notice that the noise level has decreased over 10 dB, confirming that the noise reduction techniques in the stopbands were successful. Also, notice that the stopband noise in the third segment is slightly lower than in the first segment. This is due to the narrower IF bandwidth of the third segment (300 Hz).

Making Measurements Using the Swept List Mode to Test a Device
Figure 1-57 Filter Measurements Using Linear Sweep and Swept List Mode

Using Swept List Mode


Using Linear Sweep

Making Measurements Using Limit Lines to Test a Device
Limit testing is a measurement technique that compares measurement data to constraints that you define. Depending on the results of this comparison, the analyzer will indicate if your device either passes or fails the test.
Limit testing is implemented by creating individual flat, sloping, and single-point limit lines on the analyzer display. When combined, these lines can represent the performance parameters for your device under test. The limit lines created on each measurement channel are independent of each other.
This example measurement shows you how to test a bandpass filter using the following procedures:
Figure 1-58 Connections for SAW Filter Example Measurement
Center 134 M/µ
Span 50 M/µ
Scale Ref AUTO SCALE
You may also want to select settings for the number of data points, power, averaging, and IF bandwidth.

Making Measurements Using Limit Lines to Test a Device
Cal CALIBRATE MENU RESPONSE THRU
Scale Ref AUTO SCALE
In this example procedure, the following flat limit line values are set:
Frequency Range Power Range
127 MHz to 140 MHz −27 dB to −21 dB
100 MHz to 123 MHz −200 dB to −65 dB
146 MHz to 160 MHz −200 dB to −65 dB
System LIMIT MENU LIMIT LINE LIMIT LINE ON EDIT LIMIT LINE CLEAR LIST YES
The analyzer generates a new segment that appears on the center of the display.

MIDDLE VALUE −24 x1 DELTA LIMITS 3 x1
This would correspond to a test specification of −24 ±3 dB.

Making Measurements Using Limit Lines to Test a Device
ADD STIMULUS VALUE 140 M/µ DONE LIMIT TYPE SINGLE POINT RETURN
Figure 1-59 shows the flat limit lines that you have just created with the following parameters:
Figure 1-59 Example Flat Limit Line

STIMULUS VALUE 100 M/µ
UPPER LIMIT −65 x1
LOWER LIMIT −200 x1
LIMIT TYPE FLAT LINE RETURN
ADD
STIMULUS VALUE 123 M/µ
LIMIT TYPE SINGLE POINT RETURN

Making Measurements Using Limit Lines to Test a Device
STIMULUS VALUE 146 M/µ
UPPER LIMIT −65 x1
LOWER LIMIT −200 x1
LIMIT TYPE FLAT LINE RETURN
ADD
STIMULUS VALUE 160 M/µ
LIMIT TYPE SINGLE POINT RETURN
Figure 1-60 Example Flat Limit Lines

This example procedure shows you how to make limits that test the shape factor of a SAW filter. The following limits are set:
Frequency Range Power Range
123 MHz to 125 MHz −65 dB to −26 dB
144 MHz to 146 MHz −26 dB to −65 dB

Making Measurements Using Limit Lines to Test a Device
System LIMIT MENU LIMIT LINE LIMIT LINE ON EDIT LIMIT LINE

UPPER LIMIT −65 x1 LOWER LIMIT −200 x1

UPPER LIMIT −26 x1 LOWER LIMIT −200 x1

UPPER LIMIT −26 x1 LOWER LIMIT −200 x1

UPPER LIMIT −65 x1 LOWER LIMIT −200 x1
You could use this type of limit to test the shape factor of a filter.

Making Measurements Using Limit Lines to Test a Device
Figure 1-61 Sloping Limit Lines

In this example procedure, the following limits are set:
System LIMIT MENU LIMIT LINE LIMIT LINE ON EDIT LIMIT LINE CLEAR LIST YES
Press:

UPPER LIMIT −23 x1 LOWER LIMIT −28.5 x1 DONE LIMIT TYPE SINGLE POINT RETURN
LOWER LIMIT −28.5 x1 DONE LIMIT TYPE SINGLE POINT RETURN

Making Measurements Using Limit Lines to Test a Device
Figure 1-62 Example Single Points Limit Line

This example shows you how to edit the upper limit of a limit line.
System LIMIT MENU LIMIT LINE LIMIT LINE ON EDIT LIMIT LINE
or repeatedly OR and enter the segment number followed by .
EDIT UPPER LIMIT −20 x1 DONE
System LIMIT MENU LIMIT LINE LIMIT LINE ON EDIT LIMIT LINE
or repeatedly OR and enter the segment number followed by .
Deleting Limit Segments

Making Measurements Using Limit Lines to Test a Device
System LIMIT MENU LIMIT LINE LIMIT LINE ON EDIT LIMIT LINE
Reviewing the Limit Line Segments
The limit table data that you have previously entered is shown on the analyzer display.
Activating the Limit Test
To activate the limit test and the beep fail indicator, press:
System LIMIT MENU LIMIT LINE LIMIT TEST ON BEEP FAIL ON
LIMIT LINE ON
The limit test results appear on the right side on the analyzer display. The analyzer indicates whether the filter passes or fails the defined limit test:
The limit offset functions allow you to adjust the limit lines to the frequency and output level of your device. For example, you could apply the stimulus offset feature for testing tunable filters. Or, you could apply the amplitude offset feature for testing variable attenuators, or passband ripple in filters with variable loss.
This example shows you the offset feature and the limit test failure indications that can appear on the analyzer display.
and

Making Measurements Using Limit Lines to Test a Device
System LIMIT MENU LIMIT LINE LIMIT LINE OFFSETS
STIMULUS OFFSET 3 M/µ
The analyzer beeps and a FAIL notation appears on the analyzer display, as shown in Figure 1-63.
Figure 1-63 Example Stimulus Offset of Limit Lines

STIMULUS OFFSET 0 x1
AMPLITUDE OFFSET 5 x1
The analyzer beeps and a FAIL notation appears on the analyzer display.
AMPLITUDE OFFSET 0 x1

Making Measurements Using Ripple Limits to Test a Device
Two tasks are involved in preparing for ripple testing:
This example will show you how to set up the analyzer to test ripple limits. In this example, we will be testing the pass band of a bandpass filter where the center frequency of the filter is approximately 1.8 GHz and has a bandwidth of approximately 2.9 GHz. Refer to Figure 1-64.
Figure 1-64 Bandpass Filter Being Ripple Tested

Setting Up the Analyzer to Perform the Ripple Test
This section sets up the analyzer so that a bandpass filter can be easily viewed on the analyzer display.

Making Measurements Using Ripple Limits to Test a Device
Figure 1-65 Connections for an Example Ripple Test Measurement
Center 1.8 G/n
Span 3.4 G/n
Scale Ref AUTO SCALE
You may also want to select settings for the number of data points, power, averaging, and IF bandwidth.
Cal CALIBRATE MENU RESPONSE THRU
Scale Ref AUTO SCALE

Making Measurements Using Ripple Limits to Test a Device
Figure 1-66 Filter Pass Band Before Ripple Test

Setting Up Limits for Ripple Testing
This section instructs you on setting up the ripple test parameters. You must set up the analyzer to check the DUT at the correct frequencies and compare the measured values against the maximum allowable ripple value for each frequency band. To do this, you set up individual frequency bands. You define the stop and start frequency and the maximum allowable ripple value of each frequency band. You may set up as many as 12 frequency bands for testing ripple. The frequency bands are combined in a list that is displayed while the ripple frequency bands are being edited.
In this example, we will create one ripple limit (or frequency band) that spans the entire pass band from 500 MHz to 3.0 GHz. We will also create two additional frequency bands that when merged, will span the pass band with tighter limits. Using the Ripple Edit Menu, we will create a ripple limits list on the analyzer that is similar to the following table.
Table 1-3 Ripple Limits for Ripple Test Example
1 500 MHz 3.2 GHz 2.0 dB
2 500 MHz 1.85 GHz 1.3 dB
3 1.85 GHz 3.2 GHz 1.3 dB

Making Measurements Using Ripple Limits to Test a Device
System LIMIT MENU RIPPLE LIMIT
MINIMUM FREQUENCY 500 M/µ
MAXIMUM FREQUENCY 3.2 G/n
MAXIMUM RIPPLE 2.0 x1
The network analyzer allows you to enter up to 12 frequency bands to be tested for maximum ripple.
Editing Ripple Test Limits
Once the frequency band limits for ripple testing have been created, the limits may be changed using the same menu that was used to create them. Using the edit ripple test menu, you may:
Changing Existing Frequency Band Limits
Existing frequency band limits may be changed for testing the ripple. This procedure guides you through changing the existing frequency band limits.
a.
b. The numeric key indicating the frequency band number that you are changing.
The frequency band number is located in the left column of the list of frequency bands.

Making Measurements Using Ripple Limits to Test a Device
a. and the new value to change the lower frequency of the frequency band.
b. and the new value to change the upper frequency of the frequency band.
c. and the new decibel value to change the maximum allowable ripple of the frequency band. Terminate the new decibel value with the key.
Adding Additional Frequency Bands
More frequency band limits may be added for testing the ripple. This procedure guides you through adding the more frequency band limits.
The network analyzer allows you to enter up to 12 frequency bands for maximum ripple testing.
a.
b. the numeric keys indicating the minimum frequency value of the frequency band
c. the appropriate frequency key (either , , or )
G/n M/µ k/m
a.
b. the numeric keys indicating the maximum frequency value of the frequency band
G/n M/µ k/m
c. the appropriate frequency key (either , , or )
a.
b. the decibel value of the frequency band’s maximum allowable ripple

Making Measurements Using Ripple Limits to Test a Device
Deleting Existing Frequency Bands
Frequency band limits may be deleted for testing the ripple. This procedure guides you through deleting existing frequency band limits. You may delete individual frequency bands or delete all of the frequency bands from the list.
FREQUENCY BAND 3 x1 DELETE
When this softkey is pressed, you will be asked to confirm that you want to delete all of the frequency bands from the list.
Running the Ripple Test
Once the list of ripple limits has been set up, you are ready to run the ripple test. From the Ripple Test Menu, you can:
J the absolute measured ripple value
J the margin which the measured ripple passes or fails the user-defined maximum
Starting and Stopping the Ripple Test
Once the list of ripple limits has been set up, start the ripple test by pressing
from the Ripple Test Menu until ON is displayed on the softkey. Pressing this softkey toggles the analyzer between ripple test on and ripple test off status. Figure 1-67 shows the filter pass band (with the scale changed to 1 dB/division) being ripple tested. Note that the filter fails the ripple test. The portions of the pass band trace which do not meet the test requirements are displayed in red.
RIPL TEST on OFF
ripple value

Making Measurements Using Ripple Limits to Test a Device
Figure 1-67 Filter Passband with Ripple Test Activated

As the analyzer measures the ripple, a message is displayed indicating whether the measurement passes or fails:
Displaying the Ripple Limits
After the list of ripple limits has been set up, display the ripple test limits by pressing
from the Ripple Test Menu until ON is displayed on the softkey. Pressing this softkey toggles the analyzer ripple limits display on and off. If the ripple limits are displayed and the ripple test is off, the ripple limits are displayed near the top of the graticule and are not compared with the displayed trace. However, once the ripple test is started, the ripple limits are displayed with respect to the measured trace in the following manner:
RIPL LIMIT on OFF

Making Measurements Using Ripple Limits to Test a Device
Figure 1-68 shows the filter pass band tested with the ripple limits activated. Notice that there are three sets of ripple limits shown. Also notice that the measured trace exceeds the upper ripple limit only in Frequency Band 3.
Figure 1-68 Filter Pass Band with Ripple Test and Ripple Limits Activated

Changing the Ripple Limits Line Color. The color of the lines that represent the ripple limits can be changed by:
Checking the Ripple Value
Once the ripple test has been started and is running, you may display the ripple value of each frequency band in one of two formats, the absolute format or the margin format. Both formats are described in this section.
Display
![Page 103 of the Agilent Technologies Network Analyzer (20ES) user manual — RIPL VALUE [ABSOLUTE ]](/cdn/reader/15/150a289e983abebdc4bd6f0c3204b03f577bfd829d08c0cf5bb1353a14e103d7/images/page-103.webp?w=400)
Making Measurements Using Ripple Limits to Test a Device
To display the ripple value, press . Pressing this softkey toggles between , , and . from the Ripple Test Menu until ON is displayed on the softkey. Pressing this softkey toggles the analyzer between ripple test on and ripple test off status.
RIPL VALUE [ ] RIPL VALUE [OFF] RIPL VALUE [ABSOLUTE ] RIPL VALUE [MARGIN ] RIPL TEST on OFF
When the Absolute and Margin choices are selected, the frequency band and measurement value are displayed to the right side of the pass/fail message described previously. This display is displayed in the same color as the pass/fail message.
The frequency band of the displayed value is displayed as Bn (where n = the frequency band number). The frequency band may be changed to display the value of each band. To change the displayed frequency band value, from the Ripple Test Menu press
and then use the and keys (or the numerical keypad) to select the desired frequency band.
Viewing the Ripple Value in Absolute Format
When is selected, the absolute ripple value of the selected frequency band is displayed. The absolute ripple value is the measured maximum level minus the measured minimum level within the frequency band. This value is displayed in dB.
Figure 1-69 shows the ripple test with absolute ripple value displayed for Frequency Band 1. The B1 indicates that the ripple value displayed is for Frequency Band 1. Notice that Frequency Band 1 passes the ripple test. It has an absolute ripple value of 1.675 dB while the maximum ripple value entered for Frequency Band 1 was 2.0 dB. Thus, even though the ripple test fails because of Frequency Band 3, the ripple passes in Frequency Band 1.
![Page 104 of the Agilent Technologies Network Analyzer (20ES) user manual — RIPL VALUE [MARGIN ]](/cdn/reader/15/150a289e983abebdc4bd6f0c3204b03f577bfd829d08c0cf5bb1353a14e103d7/images/page-104.webp?w=400)
Making Measurements Using Ripple Limits to Test a Device
Figure 1-69 Filter Pass Band with Absolute Ripple Value for Band 1 Activated

Viewing the Ripple Value in Margin Format
When is selected, the margin by which the ripple value passed or failed is displayed. The ripple value margin is the user-defined maximum ripple minus the absolute ripple value within the frequency band. This value is displayed in dB. A positive value is the margin by which the ripple passes the ripple test in the frequency band. A negative value is the margin by which the ripple fails the ripple test in the frequency band.

Making Measurements Using Ripple Limits to Test a Device
Figure 1-70 shows the ripple test with margin ripple value displayed for Frequency Band 2. Notice that Frequency Band 2 passes the ripple test with a margin of 0.097 dB. The plus sign (+) indicates this band passes the ripple test by the amount displayed. A minus sign (−) would indicate that the band failed by the displayed amount.
Figure 1-70 Filter Pass Band with Margin Ripple Value for Band 2 Activated


Making Measurements Using Bandwidth Limits to Test a Bandpass Filter
The bandwidth testing mode can be used to test the bandwidth of a bandpass filter. The bandwidth test finds the peak of a signal in the passband and locates a point on each side of the passband at an amplitude below the peak (that you specify during the test setup). The frequency between these two points is the bandwidth of the filter. This bandwidth is compared to minimum and maximum allowable bandwidths that you specify during the test setup.
This example shows you how to test the bandwidth of a bandpass filter. In this example, we will be testing the pass band of a bandpass filter where the center frequency of the filter is approximately 321 MHz. Refer to Figure 1-71.
Figure 1-71 Bandpass Filter Being Bandwidth Tested

When you set up the bandwidth limits to test the bandpass filter, you will first set up the analyzer to perform the bandwidth test and then you will set up bandwidth limits of the bandwidth test.
Setting Up the Analyzer to Perform the Bandwidth Test
This section sets up the analyzer so that a bandpass filter can be easily viewed on the analyzer display.

Making Measurements Using Bandwidth Limits to Test a Bandpass Filter
Figure 1-72 Connections for a Bandpass Filter Example Measurement
a. or on ET models:
b.
Center 321 M/µ
Span 200 M/µ
d.
Scale Ref AUTO SCALE
You may also want to select settings for the number of data points, power, averaging, and IF bandwidth.
Figure 1-73 Filter Pass Band Before Bandwidth Test

Preset

Making Measurements Using Bandwidth Limits to Test a Bandpass Filter
Cal CALIBRATE MENU RESPONSE THRU
Refer to Figure 1-73.
Setting Up the Bandwidth Limits
When you set up the bandwidth limits to test the bandpass filter, you will set
System LIMIT MENU BANDWIDTH LIMIT
N DB POINTS 40 x1
MINIMUM BANDWIDTH 100 M/µ
MAXIMUM BANDWIDTH 150 M/µ
After setting up the bandwidth limits, you are ready to run the bandwidth test and check the test results. For this example, we will:

Making Measurements Using Bandwidth Limits to Test a Bandpass Filter
Activating the Bandwidth Test
BW TEST on OFF
The bandwidth test continues to run until the softkey is returned to the OFF position.
The test displays a message in the upper left corner of the graticule showing that the bandwidth test is being performed and the channel on which the test is being performed. For example, BW1: indicates that the bandwidth test is being run on channel 1. See Figure 1-74.
The test also displays a message indicating whether the filter passes or fails the bandwidth test. When the filter is passing the test, the message indicates Pass. When the filter is failing the test, the failure message indicates either Wide (when the pass band is wider than the maximum bandwidth input) or Narrow (when the pass band is narrower than the minimum bandwidth input).
When the filter passes the bandwidth test, the color of the bandwidth test Pass message is green. When the filter fails the bandwidth test, the color of the bandwidth test Wide/Narrow message is red.
Figure 1-74 Filter Pass Band with Bandwidth Test Activated

Displaying the Bandwidth Markers
BW MARKER on OFF
When the bandwidth markers are displayed, a marker is placed on each side of the peak amplitude at a position equal to the N dB Points value below the peak. The markers are placed at the 40 dB points on the signal in Figure 1-75. The bandwidth markers resemble the following symbol: T

Making Measurements Using Bandwidth Limits to Test a Bandpass Filter
Figure 1-75 Bandwidth Markers Placed 40 dB Below the Bandpass Peak

Displaying the Bandwidth Value
BW DISPLAY on OFF
When this softkey is set to the ON position, the measured bandwidth value is displayed in the upper left corner of the display, to the right of the bandwidth Pass/Wide/Narrow message. This value changes as the analyzer continues measuring the bandwidth. The bandwidth value is displayed in Figure 1-76.
If the filter is failing the bandwidth test, the color of the bandwidth value is red, the same color as the failure (Wide) message of Figure 1-74. If the filter is passing the bandwidth test, the displayed bandwidth value is green (the same color as the bandwidth test Pass message).

Making Measurements Using Bandwidth Limits to Test a Bandpass Filter
Figure 1-76 Filter Pass Band with Bandwidth Value Displayed


Making Measurements Using Test Sequencing
Test sequencing allows you to automate repetitive tasks. As you make a measurement, the analyzer memorizes the keystrokes. Later you can repeat the entire sequence by pressing a single key. Because the sequence is defined with normal measurement keystrokes, you do not need additional programming expertise. Subroutines and limited decision-making increases the flexibility of test sequences. In addition, the GPIO outputs can be controlled in a test sequence, and the GPIO inputs can be tested in a sequence for conditional branching.
The test sequence function allows you to create, title, save, and execute up to six independent sequences internally. You can also save sequences to disk and transfer them from the analyzer to another analyzer or possibly to an external computer controller (so the sequence can be sent to another analyzer).
The following procedures, which are based on an actual measurement example, show you how to do the following:
Seq NEW SEQ/MODIFY SEQ
As shown in Figure 1-77, a list of instructions appear on the analyzer display to help you create or edit a sequence.

Making Measurements Using Test Sequencing
Figure 1-77 Test Sequencing Help Instructions

This choice selects sequence position #1. The default title is SEQ1 for this sequence.
Refer to "Changing the Sequence Title" on page 1-103 for information on how to modify a sequence title.
Save/Recall SELECT DISK INTERNAL MEMORY
Use the front panel knob to scroll until Preset State is highlighted on the display.


Making Measurements Using Test Sequencing
The previous keystrokes will create a displayed list as shown:
Start of Sequence RECALL PRST STATE Trans: FWD S21 (B/R) LOG MAG CENTER 134 M/u SPAN 50 M/u SCALE/DIV AUTO SCALE
Seq DONE SEQ MODIFY
To run a stored test sequence, press:
and the softkey labeled with desired sequence number.
or, press:
and the softkey labeled with the desired sequence number
To stop a sequence before it has finished, press .
Seq NEW SEQ/MODIFY SEQ

Making Measurements Using Test Sequencing
Seq DONE SEQ MODIFY
Seq NEW SEQ/MODIFY SEQ
Seq DONE SEQ MODIFY
Preset Seq NEW SEQ/MODIFY SEQ

Making Measurements Using Test Sequencing
The following list is the commands entered in "Creating a Sequence" on page 1-98. Notice that for longer sequences, only a portion of the list can appear on the screen at one time.
Start of Sequence RECALL PRST STATE Trans: FWD S21 (B/R) LOG MAG CENTER 134 M/u SPAN 50 M/u SCALE/DIV AUTO SCALE
Seq DONE SEQ MODIFY
Seq MORE CLEAR SEQUENCE
or · 75 M/µ

Making Measurements Using Test Sequencing
If you are storing sequences on a disk, you should replace the default titles (SEQ1, SEQ2, …).
Seq MORE TITLE SEQUENCE
and select the particular sequence softkey.
The analyzer shows the available title characters. The current title is displayed in the upper-left corner of the screen.
SELECT LETTER
as you stop at each character.
The analyzer cannot accept a title (file name) that is longer than eight characters. Your titles must also begin with a letter, and contain only letters and numbers.
Naming Files Generated by a Sequence
The analyzer can automatically increment the name of a file that is generated by a sequence using a loop structure. (See example "Loop counter decision making" on page 1-113.)
To access the sequence file name menu, press:
SEQUENCE FILENAMING
This menu presents two choices:
These keys show the current file name in the 2nd line of the softkey.
When titling a file for use in a loop function, you are restricted to only 2 characters in the file name due to the 6 character length of the loop counter keyword “[LOOP].” When the file is actually written, the [LOOP] keyword is expanded to only 5 ASCII characters (digits), resulting in a 7 character file name.
After entering the 2 character file name, press:

Making Measurements Using Test Sequencing
Seq MORE STORE SEQ TO DISK
and select the particular sequence softkey.
The disk drive access light should turn on briefly. When it goes out, the sequence has been saved.
For this procedure to work, the desired file must exist on the disk in the analyzer drive.
Seq MORE LOAD SEQ FROM DISK READ SEQ FILE TITLS
until the desired file name appears.
Seq MORE STORE SEQ TO DISK PURGE SEQUENCES READ SEQ FILE TITLS
until the desired file name appears.

Making Measurements Using Test Sequencing
Seq MORE PRINT SEQUENCE
and the softkey for the desired sequence.
Features That Operate Differently When Executed in a Sequence
The analyzer does not allow you to use the following keys in a sequence:
Commands That Sequencing Completes Before the Next Sequence Command Begins
The analyzer completes all operations related to the following commands before continuing with another sequence command:
*Wait 0 is the special sequencing function with a zero entered for the delay value.

Making Measurements Using Test Sequencing
Commands That Require a Clean Sweep
Many front panel commands disrupt the sweep in progress, for example, changing the channel or measurement type. When the analyzer does execute a disruptive command in a sequence, some instrument functions are inhibited until a complete sweep is taken. This applies to the following functions:
Forward Stepping in Edit Mode
In the sequence modify mode, you can step through the selected sequence list, where the analyzer executes each step using the .
A title may contain non-printable or special ASCII characters if you download it from an external controller. A non-printable character is represented on the display as π.
A sequence may contain up to 2 kbytes of instructions. Typically, this is around 200 sequence command lines. To estimate a sequence's size (in kbytes), use the following guidelines.
Table 1-4 Guidelines for Determining the Size of a Sequence
Type of Command Size in Bytes
Title string character 1
Active entry command 1 per digit
Embedding the Value of the Loop Counter in a Title
You can append a sequentially increasing or decreasing numeric value to the title of stored data by placing a command after the title string. (You must limit the title to three characters if you will use it as a disk file name. The three-character title and five-digit loop counter number reach the eight-character limit for disk file names.) This feature is useful in data logging applications.
Display MORE TITLE MORE LOOP COUNTER
You can define a sequence to run automatically when you apply power to the analyzer. To make an autostarting sequence, create a sequence in position six (SEQ6) and title it “AUTO”. To stop an autostarting sequence, press . To stop an autostarting sequence from engaging at power on, you must clear it from memory or rename it.

Making Measurements Using Test Sequencing
The softkey, located in the Sequencing menu, activates a feature that allows the sequence to branch off to another sequence, then return to the original sequence. For example, you could perform an amplifier measurement in the following manner:
The instrument’s parallel port can be used in two different modes. By pressing and then toggling the softkey, you can select either the mode or the mode.
Local PARALLEL [ ] [COPY] [GPIO]
The GPIO mode switches the parallel port into a “general purpose input/output” port.
In this mode, the port can be connected to test fixtures, power supplies, and other peripheral equipment that the analyzer can interact with through test sequencing.
The TESTSET I/O interconnect on the rear panel was originally intended for use solely with the HP/Agilent 85046A/B and HP/Agilent 85047B external S-parameter test sets. Since the introduction of the 8753D, a network analyzer with an internal test set, this test set I/O port has become a general purpose control port for a variety of external devices, such as the K36 or K39 test adapters and Option 014 configurations. Refer Table 1-5 on page 1-111 for the definition of each pin of the test set I/O connector.
This connector, with the limit output, TTL OUT and TESTSET I/O outputs can also be used with part handlers to provide control interface.
Gosub Sequence Command · The GPIO Mode

Making Measurements Using Test Sequencing
The TESTSET I/O bits are set using the and keys under the keys. The values of the outputs (pins 11, 22, and 23) are described in Table 1-5. The value changes with the test port, so if the external control is required for both test port directions, the settings must be made under both
Seq TTL I/O TTL OUT
and . This capability can be used to set different external conditions in a test requiring changes between the forward and reverse measurements, as might be needed in a high power test, for example.
This menu can be accessed by pressing in the Sequencing menu.
TTL Output for Controlling Peripherals Eight TTL compatible output lines can be used for controlling equipment connected to the parallel port. By pressing , you will access the following softkeys that control the individual output bits. Refer to Figure 1-78 for output bus pin locations.
lets you input a number (0 to 255) in base 10 and outputs it to the bus as binary.
lets you set a single bit (0 − 7) to high on the output bus.
CLEAR BIT
lets you set a single bit (0 − 7) to low on the output bus.
TTL Input Decision Making Five TTL compatible input lines can be used for decision making in test sequencing. For example, if a test fixture is connected to the parallel port and has a micro switch that needs to be activated in order to proceed with a measurement, you can construct your test sequence so that it checks the TTL state of the input line corresponding to the switch. Depending on whether the line is high or low, you can jump to another sequence. To access these decision making functions, press . Refer to Figure 1-78 for input bus pin locations.
lets you select the single bit (0 − 4) that the sequence will be looking for.
lets you jump to another sequence if the single input bit you selected is in a high state.
lets you jump to another sequence if the single input bit you selected is in a low state.
Pin assignments:
TTL I/O Menu · Seq TTL I/O · Seq TTL I/O

Making Measurements Using Test Sequencing
Electrical specifications for TTL high:
Electrical specifications for TTL low:
Figure 1-78 Parallel Port Input and Output Bus Pin Locations in GPIO Mode

Making Measurements Using Test Sequencing
Test Set Interconnect Control
Figure 1-79 Test Set Interconnect Pin Designations

Control of the external switch (8762B Option T24) can be done through the test set interface on the rear panel of the analyzer.
Pin 22 (TTL 1) on the TEST SET-I/O INTERCONNECT connector is a TTL line that changes from TTL high to TTL low when changing TTL I/O FWD from 7 to 6. Refer to Figure 1-79. To change from 7 to 6, press the following sequence:
TTL I/O TTL OUT TESTSET I/O FWD 6 x1
Changing the switch state back to the standard mode requires a 7 to be entered in the “TESTSET I/O FWD.”
Pin 1 on the external switch must be grounded. It can be grounded to:
Pin C (common) on the external switch (8762B Option T24) must be connected to the test set interface pin 14 (+22 volt line). Refer to Figure 1-79.
Pin 2 on the external switch, connects to pin 22 (TTL 1) on the test set interface.
The TTL I/O can control both of the external RF switches. Both must be cascaded in parallel together. Changing the TTL I/O FWD from 7 to 6 will change the external RF switch state. This changes the measurement capability from the network analyzer to the external test measurement device. The TTL I/O FWD when changed from 7 to 6 will reverse the process.

Making Measurements Using Test Sequencing
Table 1-5 Test Set Interconnect Pin Designation
Pin Number Pin Description
Pin 1 No Connection (NC)
Pin 3 Same as Test Sequence (TTL OUT) output BNC connector
Pin 8 Hi-forward/Low-reverse. Follows the test port indicator.
Pin 10 Lstarttrig: Not used. Do not connect anything to this pin.
Pin 11 TESTSET I/O Bit 2 (most significant bit). +5 V when TESTSET I/O has values of 4, 5, 6, or 7. Otherwise, bit is TTL low.
Pin 14 +22 Volts
Pin 17 Same as Limit Test output BNC connector
Pin 22 TESTSET I/O Bit 0 (least significant bit). +5 V when TESTSET I/O has values of 1, 3, 5, or 7. Otherwise, bit is TTL low.
Pin 23 TESTSET I/O Bit 1 (middle bit). +5 V when TESTSET I/O has values of 2, 3, 6, or 7. Otherwise, bit is TTL low.
Pin 24 Lremtrig: TTL low when TEST- SET I/O pins are valid. This bit can be used to latch these values.

Making Measurements Using Test Sequencing
TTL Out Menu The softkey provides access to the TTL out menu. This menu allows you to choose between the following output parameters of the TTL output signal:
TTL OUT HIGH
TTL OUT LOW
END SWEEP HIGH PULSE
END SWEEP LOW PULSE
The TTL output signals are sent to the sequencing BNC rear panel output.
Sequencing Special Functions Menu
This menu is accessed by pressing the softkey in the Sequencing menu. This menu provides access to the decision making menu and the more special functions menu. It also contains the peripheral GPIB address and titling, wait x, pause, and marker → CW functions.
Sequence Decision Making Menu
This menu is accessed by pressing the softkey in the Sequencing Special Functions menu.
Decision making functions are explained in more detail in the following section. These functions check a condition and jump to a specified sequence if the condition is true. The sequence called must be in memory. A sequence call is a one-way jump. A sequence can jump to itself, or to any of the other five sequences currently in memory. Use of these features is explained under the specific softkey descriptions.
Decision making functions jump to a softkey location, not to a specific sequence title Limit test, loop counter, and do sequence commands jump to any sequence residing in the specified sequence position (1 through 6). These commands do not jump to a specific sequence title. Whatever sequence is in the selected softkey position, will run when these commands are executed.
Having a sequence jump to itself A decision making command can jump to the sequence it is in. When this occurs, the sequence starts over and all commands in the sequence are repeated. This is used a great deal in conjunction with loop counter commands. See the loop counter description that follows.
TTL input decision making TTL input from a peripheral connected to the parallel port (in the GPIO mode) can be used in a decision making function. Refer to "The GPIO Mode" on page 1-107.
Limit test decision making A sequence can jump to another sequence or start over depending on the result of a limit test. When entered into a sequence, the
and commands require you to enter the destination sequence.

Making Measurements Using Test Sequencing
Loop counter decision making The analyzer has a numeric register called a loop counter. The value of this register can be set by a sequence, and it can be incriminated or decremented each time a sequence repeats itself. The decision making commands
and jump to another sequence if the stated condition is true. When entered into the sequence, these commands require you to enter the destination sequence. Either command can jump to another sequence, or restart the current sequence.
IF LOOP COUNTER = 0 IF LOOP COUNTER <> 0
As explained in "Embedding the Value of the Loop Counter in a Title" on page 1-106, the loop counter value can be appended to a title. This allows customized titles for data printouts or for data files saved to disk.

Making Measurements Using Test Sequencing to Test a Device
Test sequencing allows you to automate repetitive tasks. As you make a measurement, the analyzer memorizes the keystrokes. Later you can repeat the entire sequence by pressing a single key.
This section contains the following example sequences:
By cascading test sequences, you can create subprograms for a larger test sequence. You can also cascade sequences to extend the length of test sequences to greater than 200 lines.
In this example, you are shown two sequences that have been cascaded. You can do this by having the last command in sequence 1 call sequence position 2, regardless of the sequence title. Because sequences are identified by position, not title, the call operation will always go to the sequence loaded into the given position.
Seq NEW SEQ/MODIFY SEQ SEQUENCE 1 SEQ1
Center 134 M/µ
Span 50 M/µ
Seq DO SEQUENCE SEQUENCE 2
NEW SEQ/MODIFY SEQ SEQUENCE 2 SEQ2
or on ET models:
Seq DONE SEQ MODIFY
Format LOG MAG · Scale Ref AUTOSCALE

Making Measurements Using Test Sequencing to Test a Device
The following sequences will be created:
SEQUENCE SEQ1 Start of Sequence CENTER 134 M/u SPAN 50 M/u DO SEQUENCE SEQUENCE 2
SEQUENCE SEQ2 Start of Sequence Trans:FWD S21 (B/R) LOG MAG SCALE/DIV AUTO SCALE
You can extend this process of calling the next sequence from the last line of the present sequence to 6 internal sequences, or an unlimited number of externally stored sequences.
Preset SEQUENCE 1 SEQ1
This example shows you the basic steps necessary for constructing a looping structure within a test sequence. A typical application of this loop counter structure is for repeating a specific measurement as you step through a series of CW frequencies or dc bias levels. For an example application, see "Fixed IF Mixer Measurements" on page 2-27.
Seq NEW SEQ/MODIFY SEQ SEQUENCE 1 SEQ1
LOOP COUNTER 10 x1
Seq DO SEQUENCE SEQUENCE 2
This will create a displayed list as shown:
SEQUENCE LOOP 1 Start of Sequence LOOP COUNTER 10x1 DO SEQUENCE SEQUENCE 2

Making Measurements Using Test Sequencing to Test a Device
To create a second sequence that will perform a desired measurement function, decrement the loop counter, and call itself until the loop counter value is equal to zero, press:

This will create a displayed list as shown:
SEQUENCE LOOP 2 Start of Sequence Trans:FWD S21 (B/R) SCALE/DIV AUTO SCALE MKR Fctn SEARCH MAX DECR LOOP COUNTER IF LOOP COUNTER 0 THEN DO SEQUENCE 2
To run the loop sequence, press:
Preset SEQUENCE 1 SEQ1
This example shows how to increment the names of files that are generated by a sequence with a loop structure.


Making Measurements Using Test Sequencing to Test a Device
Save/Recall FILE UTILITIES SEQUENCE FILE NAMING
D T LOOP COUNTER DONE
PLOT NAME PLOTFILE ERASE TITLE
P L LOOP COUNTER DONE RETURN
Sweep Setup TRIGGER MENU SINGLE
Seq SPECIAL FUNCTIONS DECISION MAKING
SEQUENCE 2 SEQ 2
Seq DONE SEQ MODIFY
This will create the following displayed lists:
Start of Sequence LOOP COUNTER
7 x1 INTERNAL DISK DATA ONLY
SEQUENCE 2
Start of Sequence FILE NAME
PL[LOOP] SINGLE SAVE FILE 0 PLOT DECR LOOP COUNTER IF LOOP COUNTER 0 THEN DO
Sequence 1 initializes the loop counter and calls sequence 2. Sequence 2 repeats until the loop counter reaches 0. For each loop, it takes a single sweep, saves the data file and plots the display.
DT000001.D1
Save/Recall SAVE STATE · Copy PLOT · through

Making Measurements Using Test Sequencing to Test a Device
PL00001.FP
To run the sequence, press:
Preset SEQUENCE 1 SEQ 1
This measurement example shows you how to create a sequence that will branch the sequence according to the outcome of a limit test. Refer to "Using Limit Lines to Test a Device" on page 1-72, for a procedure that shows you how to create a limit test.
For this example, you must have already saved the following in register 1:
Seq NEW SEQ MODIFY SEQUENCE 1 SEQ1
Save/Recall RECALL KEYS RECALLKEYS MENU RECALL REG1
Seq SPECIAL FUNCTIONS DECISION MAKING
IF LIMIT TEST FAIL SEQUENCE 3 SEQ3
Seq DONE SEQ MODIFY
This will create a displayed list for sequence 1, as shown:
Start of Sequence RECALL REG 1 IF LIMIT TEST PASS THEN DO SEQUENCE 2 IF LIMIT TEST FAIL THEN DO SEQUENCE 3
Seq NEW SEQ MODIFY SEQUENCE 2 SEQ2
Save/Recall SELECT DISK INTERNAL DISK RETURN
Seq DONE SEQ MODIFY
through

Making Measurements Using Test Sequencing to Test a Device
This will create a displayed list for sequence 2, as shown:
Start of Sequence INTERNAL DISK DATA ARRAY ON FILENAME FILE 0 SAVE FILE
Seq NEW SEQ/MODIFY SEQ SEQUENCE 3 SEQ3
T U N E D E V I C E DONE
Seq SPECIAL FUNCTIONS PAUSE RETURN
Seq DONE SEQ MODIFY
This will create a displayed list for sequence 3, as shown:
Start of Sequence TITLE TUNE DEVICE SEQUENCE PAUSE DO SEQUENCE SEQUENCE 1
You will see the “tune device” prompt in the title area or the sequence will pause until you press the key.
Display MORE TITLE

Making Measurements Using Test Sequencing to Test a Device


Making Mixer Measurements (Option 089 Only) Using This Chapter
This chapter contains the following:
— Conversion loss using the frequency offset mode
— High dynamic range swept RF/IF conversion loss
— Fixed IF measurements
— Group delay measurements
— Amplitude and phase tracking
— Conversion compression using the frequency offset mode
— Isolation Measurements · — SWR/Return Loss

Making Mixer Measurements (Option 089 Only)
The analyzer is capable of measuring the following mixer (frequency converter) parameters:
Figure 2-1 Mixer Parameters

LO Local Oscillator. LO is normally provided by an external source or
internally generated by the frequency converter.
IF Intermediate Frequency. IF is usually the mixer’s output signal.
RF Radio Frequency. RF is usually the mixer’s input signal.
Mixer Measurement Capabilities

Making Mixer Measurements (Option 089 Only) Measurement Considerations
In mixer transmission measurements, you have RF and LO inputs and an IF output. Also emanating from the IF port are several other mixing products of the RF and LO signals. In mixer measurements, leakage signals from one mixer port propagate and appear at the other two mixer ports. These unwanted mixing products or leakage signals can cause distortion by mixing with a harmonic of the analyzer’s first down-conversion stage.
To ensure successful mixer measurements, the following measurement challenges must be taken into consideration:
J “Minimizing Source and Load Mismatches”
J "Reducing the Effect of Spurious Responses" on page 2-5
J "Eliminating Unwanted Mixing and Leakage Signals" on page 2-6
J "How RF and IF Are Defined" on page 2-7
J "Frequency Offset Mode Operation" on page 2-9
J "LO Frequency Accuracy and Stability" on page 2-10
J "Differences Between Internal and External R Channel Inputs" on page 2-10
J "Power Meter Calibration" on page 2-12
When characterizing linear devices, you can use vector accuracy enhancement to mathematically remove all systematic errors, including source and load mismatches, from your measurement. This is difficult when the device you are characterizing is a mixer where the input and output signals are at different frequencies. Therefore, source and load mismatches are not corrected for and will add to overall measurement uncertainty.
You should place attenuators at all of the test ports to reduce the measurement errors associated with the interaction between mixer port matches and system port matches. To avoid overdriving the receiver, you should give extra care to selecting the attenuator located at the mixer's IF port to avoid overdriving the receiver. For best results, you should choose the attenuator value so that the power incident on the analyzer R channel input is less than −10 dBm and greater than −35 dBm.

Making Mixer Measurements (Option 089 Only)
Figure 2-2 Conversion Loss versus Output Frequency without Attenuators at Mixer Ports

Figure 2-3 Example of Conversion Loss versus Output Frequency with Attenuation at All Mixer Ports

By choosing test frequencies (frequency list mode), you can reduce the effect of spurious responses on measurements by avoiding frequencies that produce IF signal path distortion.
Measurement Considerations

Making Mixer Measurements (Option 089 Only) Measurement Considerations
By placing filters between the mixer’s IF port and the receiver’s input port, you can eliminate unwanted mixing and leakage signals from entering the analyzer’s receiver. Filtering is required in both fixed and broadband measurements. Therefore, when configuring broad-band (swept) measurements, you may need to trade some measurement bandwidth for the ability to more selectively filter signals entering the analyzer receiver.
Figure 2-4 Example of Conversion Loss versus Output Frequency without Correct IF Signal Path Filtering


Making Mixer Measurements (Option 089 Only)
Figure 2-5 Example of Conversion Loss versus Output Frequency with Correct IF Signal Path Filtering and Attenuation at All Mixer Ports
When you choose between and in the frequency offset menus, the analyzer determines which direction the internal source must sweep in order to achieve the requested IF frequency. For example, to measure the lower sideband of a mixer, where the RF signal is below the LO ( ), the internal source must sweep backwards. See the examples in Figure 2-6.
RF < LO
Measurement Considerations

Making Mixer Measurements (Option 089 Only) Measurement Considerations
Figure 2-6 Examples of Up Converters and Down Converters
In standard mixer measurements, the input of the mixer is always connected to the analyzer’s RF source, and the output of the mixer always produces the IF frequencies that are received by the analyzer’s receiver.
However, the ports labeled RF and IF on most mixers are not consistently connected to the analyzer’s source and receiver ports, respectively. These mixer ports are switched, depending on whether a down converter or an up converter measurement is being performed.
It is important to keep in mind that in the setup diagrams of the frequency offset mode, the analyzer’s source and receiver ports are labeled according to the mixer port to which they are connected.
Because the RF frequency can be greater or less than the set LO frequency in this type of measurement, you can select either or .

Making Mixer Measurements (Option 089 Only)
Figure 2-7 Down Converter Port Connections

Because the RF frequency can be greater or less than the set LO frequency in this type of measurement, you can select either or .
Figure 2-8 Up Converter Port Connections

This mode of operation allows you to offset the analyzer's source by a fixed value, above or below the analyzer's receiver. That is, this allows you to use a device input frequency range that is different from the receiver input frequency range.
Measurement Considerations

Making Mixer Measurements (Option 089 Only) Measurement Considerations
Frequency offset measurements do not begin until all of the frequency offset mode parameters are set. These include the following:
The LO frequency for frequency offset mode must be set to the same value as the external LO source. The offset frequency between the analyzer source and receiver will be set to this value.
For a single-sideband mixer measurement, the RF source can be offset in frequency from the input receiver frequency, allowing for a swept RF stimulus over one frequency range and measurement of the IF response over another (in this case the output IF).
When frequency offset mode operation begins, the receiver is tuned to the entered IF signal frequencies and then offsets the source frequency required to produce the IF. Therefore, since it is the analyzer receiver that controls the source, it is only necessary to set the start and stop frequencies from the receiver.
The analyzer source is phaselocked to its receiver through a reference loop. In the frequency offset mode, the mixer under test is inserted in this loop. To ensure that the analyzer phaselocks correctly, it is important that you use an LO source that has frequency accuracy better than ±16 MHz and residual FM < 20 kHz RMS.
Due to internal losses in the analyzer's test set, the power measured internally at the R channel is 16 dB lower than that of the source. To compensate for these losses, the traces associated with the R channel have been offset 16 dB higher. As a result, power measured directly at the R channel via the R CHANNEL IN port will appear to be 16 dB higher than its actual value. If power meter calibration is not used, this offset in power must be accounted for with a receiver calibration before performing measurements.
The following steps can be performed to observe this offset in power:
PWR RANGE MAN 0 x1
Power

Making Mixer Measurements (Option 089 Only)
Setting the power range to manual prevents the internal source attenuator from switching when changing power levels. If you choose a different power range, the R channel offset compensation and R channel measurement changes by the amount of the attenuator setting.
Power −10 x1
Figure 2-9 R Channel External Connection

FREQ OFFS MENU
FREQ OFFS ON
Since the LO (offset) frequency is still set to the default value of 0 Hz, the analyzer will operate normally.
R
Observe the 13 to 16 dB offset in measured power. The actual input power level to the R channel input must be 0 dBm or less, −10 dBm typical, to avoid receiver saturation effects. The minimum signal level must be greater than −35 dBm to provide sufficient signal for operation of the phaselock loop.
Measurement Considerations · System · Meas

Making Mixer Measurements (Option 089 Only) Measurement Considerations
RECEIVER CAL −10 x1
Once completed, the R channel should display the reference power (−10 dBm in this example).
Mixer transmission measurements are generally configured as follows:
measured output power (Watts) / set input power (Watts)
measured output power (dBm) − set input power (dBm)
For this reason, the set input power must be accurately controlled in order to ensure measurement accuracy.
The amplitude variation of the analyzer is specified at ±1 dB over any given source frequency. This may give a maximum 2 dB error for a mixer transmission test setup: ±1 dB for the source over the IF range during measurement and ±1 dB over the RF range during measurement.
Higher measurement accuracy may be obtained through the use of power meter calibration. You can use power meter calibration to correct for power offsets, losses, and flatness variations occurring between the analyzer source and the input to the mixer under test. Refer to the power meter documentation for its calibration procedures.
Cal

Making Mixer Measurements (Option 089 Only) Conversion Loss Using the Frequency Offset Mode
Conversion loss is the measure of efficiency of a mixer. It is the ratio of side-band IF power to RF signal power, and is usually expressed in dB. The mixer translates the incoming signal, (RF), to a replica, (IF), displaced in frequency by the local oscillator, (LO). Frequency translation is characterized by a loss in signal amplitude and the generation of additional sidebands. For a given translation, two equal output signals are expected, a lower sideband and an upper sideband.
Figure 2-10 An Example Spectrum of RF, LO, and IF Signals Present in a
The following procedure describes the R channel swept IF frequency conversion loss measurement of a broadband component mixer with power meter calibration. For this example, we will use the following example settings. For your measurement, you will need to use settings specific to your measurement.
Settings Used for this Example
Conversion Loss Measurement

Making Mixer Measurements (Option 089 Only) Conversion Loss Using the Frequency Offset Mode
Figure 2-11 Connections for R Channel and Source Calibration

System INSTRUMENT MODE FREQ OFFS MENU
Start 650 M/µ Stop 950 M/µ
If the LO frequency is not set to 0 Hz, press:
LO FREQUENCY 0 x1
Meas INPUT PORTS R
The measurement trace is shown on the display.
Local SYSTEM CONTROLLER

Making Mixer Measurements (Option 089 Only) Conversion Loss Using the Frequency Offset Mode
(where aa is the GPIB address of the power meter)
a. Select the remote interface command set on the power meter by
pressing the System Inputs key and the following softkeys: Remote Interface, Command Set, SCPI
b. Skip step 4 and continue at step 5.
The power sensor factors are automatically read by the analyzer.
a. Choose the 438A/437 selection on the network analyzer by pressing:
b. Select the remote interface command set on the power meter by
pressing the System Inputs key and the following softkeys: Remote Interface, Command Set, 437B
Cal PWRMTR CAL LOSS/SENSR LISTS CAL FACTOR SENSOR A
fff M/µ CAL FACTOR
nnn x1 DONE DONE
Local SET ADDRESSES · until it reads

Making Mixer Measurements (Option 089 Only) Conversion Loss Using the Frequency Offset Mode

System INSTRUMENT MODE FREQ OFFS MENU
Start 100 M/µ Stop 350 M/µ
Cal CALIBRATE MENU RECEIVER CAL 0 x1 TAKE RCVR CAL SWEEP
J The low pass filter is required to limit the range of frequencies passed into the R
channel input port. The filter is selected to pass the IF frequencies for the measurement but prevent the LO feedthrough and unwanted mixer products from confusing the phase-lock loop operation.
J A pad is used to isolate the filter and improve the IF port match for the mixer.
Once completed, the display should read 0 dBm (−10 dBm for 8722ES).
Save/Recall SAVE STATE

Making Mixer Measurements (Option 089 Only) Conversion Loss Using the Frequency Offset Mode
Figure 2-12 R-Channel Mixer Measurement Equipment Setup

The analyzer is now displaying the conversion loss of the mixer calibrated with power meter accuracy.
System INSTRUMENT MODE FREQ OFFS MENU
LO FREQUENCY 1000 M/µ
The LO menu is used to set only the LO CW frequency. All other settings apply when using the HP/Agilent 8625A external source.

Making Mixer Measurements (Option 089 Only) Conversion Loss Using the Frequency Offset Mode
The measurement setup diagram is shown in Figure 2-14. Note the RF frequency values are shown in this illustration.
Figure 2-13 Diagram of Measurement Frequencies
Figure 2-14 Measurement Setup from Display


Making Mixer Measurements (Option 089 Only) Conversion Loss Using the Frequency Offset Mode
Figure 2-15 Conversion Loss Example Measurement

In this measurement, you set the input power and measured the output power. Figure 2-15 shows the absolute loss through the mixer versus mixer output frequency. If the mixer under test contained built-in amplification, then the measurement results would have shown conversion gain.
Scale Ref AUTOSCALE

Making Mixer Measurements (Option 089 Only) High Dynamic Range Swept RF/IF Conversion Loss

The frequency offset mode enables the testing of high dynamic range frequency converters (mixers), by tuning the analyzer’s high dynamic range receiver above or below its source, by a fixed offset. This capability allows the complete measurement of both pass and reject band mixer characteristics.
The analyzer has a 35 dB dynamic range limitation on measurements made directly with its R (phaselock) channel. For this reason, the measurement of high dynamic range mixing devices (such as mixers with built in amplification and filtering) with greater than 35 dB dynamic range must be made on either the analyzer’s A or B channel, with a reference mixer providing input to the analyzer’s R channel for phaselock.
This example describes the swept IF conversion loss measurement of a mixer and filter. The output filtering demonstrates the analyzer’s ability to make high dynamic range measurements.
Set the following analyzer parameters:


Making Mixer Measurements (Option 089 Only) High Dynamic Range Swept RF/IF Conversion Loss
Figure 2-16 Connections for Power Meter Calibration



Cal PWRMTR CAL LOSS/SENSR LISTS CAL FACTOR SENSOR A

Local

Making Mixer Measurements (Option 089 Only) High Dynamic Range Swept RF/IF Conversion Loss
Perform a Receiver Calibration Over the IF Range
Figure 2-17 Connections for Receiver Calibration

Meas INPUT PORTS B
Cal CALIBRATE MENU RECEIVER CAL −5 x1 −10 x1 TAKE RCVR CAL SWEEP
Once completed, the analyzer should display −5 dBm (−10 dBm, 8722ES).
You can find the RF frequency range by using a simple calculation or using the mixer measurement diagram on the analyzer display.
Add the LO frequency to the IF frequency start and stop values. If using an LO frequency of 1500 MHz, the RF start frequency would be 1.6 GHz (1500 + 100 MHz) and the stop frequency would be 2.5 GHz (1500 + 1000 MHz).
Using the Calculation

Making Mixer Measurements (Option 089 Only) High Dynamic Range Swept RF/IF Conversion Loss
Using the Mixer Measurement Diagram
While the analyzer is still set to the IF frequency range, press:
System INSTRUMENT MODE FREQ OFFS MENU LO MENU FREQUENCY:CW 1500 M/µ RETURN RETURN DOWN CONVERTER RF > LO
Press .
Start 1.6 G/n Stop 2.5 G/n
Figure 2-18 Connections for Power Meter Calibration



Making Mixer Measurements (Option 089 Only) High Dynamic Range Swept RF/IF Conversion Loss
Start 100 M/µ Stop 1
G/n

Making Mixer Measurements (Option 089 Only) High Dynamic Range Swept RF/IF Conversion Loss
Figure 2-19 Connections for a High Dynamic Range Swept IF Conversion Loss

System INSTRUMENT MODE FREQ OFFS MENU
LO FREQUENCY 1500 M/µ
RETURN DOWN CONVERTER RF > LO FREQ OFFS ON
In this low-side LO, down converter measurement, the analyzer’s source frequency range will be offset higher than the receiver frequency range. The source frequency range can be determined from the following equation:
receiver frequency range (100 – 1000 MHz) + LO frequency (1500 MHz) = 1.6 − 2.5 GHz
VIEW MEASURE Scale Ref AUTOSCALE
Figure 2-20 shows the conversion loss of this low-side LO mixer with output filtering. Notice that the dynamic range from the pass band to the noise floor is well above the dynamic range limit of the R Channel. If the mixer under test also contained amplification, then this dynamic range would have been even greater due to the conversion gain of the mixer.
Measurement

Making Mixer Measurements (Option 089 Only) High Dynamic Range Swept RF/IF Conversion Loss
Figure 2-20 Example of Swept IF Conversion Loss Measurement


Making Mixer Measurements (Option 089 Only)
Fixed IF Mixer Measurements
A fixed IF can be produced by using both a swept RF and LO that are offset by a certain frequency. With proper filtering, only this offset frequency will be present at the IF port of the mixer.
This measurement requires two external RF sources as stimuli. Figure 2-22 shows the hardware configuration for the fixed IF conversion loss measurement. This example measurement procedure uses the analyzer’s test sequence function for automatically controlling the two external synthesizers (with SCPI commands), and making a conversion loss measurement in tuned receiver mode. The test sequence function is an instrument automation feature internal to the analyzer. For more information on the test sequence function, refer to "Using Test Sequencing to Test a Device" on page 1-114.
The analyzer’s tuned receiver mode allows you to tune its receiver to an arbitrary frequency and measure signal power. This is only possible if the signal you wish to analyze is at an exact known frequency. Therefore, the RF and LO sources must be synthesized and synchronized with the analyzer’s time base.
Since the analyzer is not phaselocking in this configuration, you can use it to measure conversion loss of a microwave mixer with an RF frequency range output.
Tuned receiver mode also increases dynamic range. Broadband techniques like diode detection have a high noise floor, while narrow band techniques like tuned receivers are much less susceptible to noise.
The following sequence initializes and calibrates the network analyzer. It then initializes the two external sources prior to measurement. This sequence includes:

Making Mixer Measurements (Option 089 Only) Fixed IF Mixer Measurements
Figure 2-21 Connections for a Response Calibration

Seq NEW SEQ/MODIFY SEQ SEQUENCE 1 SEQ1
Save/Recall SELECT DISK INTERNAL MEMORY
(Select the preset state.)
Presetting the Instrument

Making Mixer Measurements (Option 089 Only)
Fixed IF Mixer Measurements
Putting the Analyzer into Tuned Receiver Mode
System INSTRUMENT MODE TUNED RECEIVER
Setting Up a Frequency List Sweep of 26 Points

Performing a Response Calibration
Meas INPUT PORTS B
Display MORE TITLE ERASE TITLE


Prompting the User to Connect a Mixer to the Test Setup
Display MORE TITLE ERASE TITLE
PAUSE
Local SYSTEM CONTROLLER · Seq SPECIAL FUNCTIONS

Making Mixer Measurements (Option 089 Only) Fixed IF Mixer Measurements
Initializing a Loop Counter Value to 26

Addressing and Configuring the Two Sources
Display MORE TITLE ERASE TITLE



Calling the Next Measurement Sequence

Press and the analyzer will display the following sequence commands:
SEQUENCE SEQ1 Start of Sequence RECALL PRST STATE SYSTEM CONTROLLER

Making Mixer Measurements (Option 089 Only)
Fixed IF Mixer Measurements
TUNED RECEIVER EDIT LIST ADD CW FREQ 100M/u NUMBER OF POINTS 26x1 DONE DONE LIST FREQ B TITLE POW:LEV 6DBM PERIPHERAL HPIB ADDR 19x1 TITLE TO PERIPHERAL TITLE FREQ:MODE CW;CW 100MHZ TITLE TO PERIPHERAL CALIBRATE: RESPONSE CAL STANDARD DONE CAL CLASS TITLE CONNECT MIXER PAUSE LOOP COUNTER 26x1 SCALE/DIV 2 x1 REFERENCE POSITION 0 x1 REFERENCE VALUE −20x1 MANUAL TRG ON POINT TITLE FREQ:MODE CW;CW 500MHZ;:FREQ:CW:STEP 100MHZ TITLE TO PERIPHERAL TITLE POW:LEV 13DBM PERIPHERAL HPIB ADDR 21x1 TITLE TO PERIPHERAL TITLE FREQ:MODE CW;CW 600MHZ;:FREQ:CW:STEP 100MHZ TITLE TO PERIPHERAL DO SEQUENCE SEQUENCE 2

Making Mixer Measurements (Option 089 Only) Fixed IF Mixer Measurements
The following sequence makes a series of measurements until all 26 CW measurements are made and the loop counter value is equal to zero. This sequence includes:

Seq SPECIAL FUNCTIONS WAIT x .1 x1
Sweep Setup TRIGGER MENU MANUAL TRG ON POINT
Incrementing the Source Frequencies
Display MORE TITLE ERASE TITLE

Decrementing the Loop Counter
DECISION MAKING DECR LOOP COUNTER IF LOOP COUNTER<>0 SEQUENCE 2 SEQ2
Display MORE TITLE ERASE TITLE
Seq DONE SEQ MODIFY
Taking Data · Labeling the Screen

Making Mixer Measurements (Option 089 Only)
Fixed IF Mixer Measurements
Seq NEW SEQ/MODIFY SEQ SEQUENCE 2 SEQ2
Press and the analyzer will display the following sequence commands:
SEQUENCE SEQ2 Start of Sequence WAIT x 1 x1 MANUAL TRG ON POINT TITLE FREQ:CW UP PERIPHERAL HPIB ADDR 19x1 TITLE TO PERIPHERAL PERIPHERAL HPIB ADDR 21x1 TITLE TO PERIPHERAL DECR LOOP COUNTER IF LOOP COUNTER 0 THEN DO SEQUENCE 2 TITLE MEASUREMENT COMPLETED
Seq DONE SEQ MODIFY DO SEQUENCE SEQUENCE 2 SEQ2
When the prompt CONNECT MIXER appears, connect the equipment as shown in Figure 2-22.
Figure 2-22 Connections for a Conversion Loss Using Tuned Receiver Mode


Making Mixer Measurements (Option 089 Only) Fixed IF Mixer Measurements
When the sequences are finished you should have a result as shown in Figure 2-23.
Figure 2-23 Example Fixed IF Mixer Measurement

The displayed trace represents the conversion loss of the mixer at 26 points. Each point corresponds to one of the 26 different sets of RF and LO frequencies that were used to create the same fixed IF frequency.

Making Mixer Measurements (Option 089 Only)
Phase or Group Delay Measurements
For information on group delay principles, refer to "Setting the Electrical Delay" on page 1-37.
When you are making linear measurements, you must provide a reference for determining phase by splitting the RF source power and send part of the signal into the reference channel. (This does not work for frequency offset measurements, since the source and receiver are functioning at different frequencies.)
To provide a reference signal for the phase measurement, you need a second mixer. This mixer is driven by the same RF and LO signals that are used to drive the mixer under test. The IF output from the reference mixer is applied to the reference (R) channel of the analyzer.
Group delay is the rate of change of phase through a device with respect to frequency (dφ/dω). Traditionally, group delay has been used to describe the propagation delay (τg), and deviation from linear phase through a linear device. However, this parameter also contains valuable information about transmission delay and distortion through a non-linear device such as a mixer or frequency converter. For example, flat group delay corresponds to low modulation distortion (that is, carrier and sidebands propagate at the same rate).
Phase linearity and group delay are both measurements of the distortion of a transmitted signal. Both measure the non-linearity of a device’s phase response with respect to frequency.
In standard vector error-correction, a thru (delay=0) is used as a calibration standard. The solution to this problem is to use a calibration mixer with very small group delay as the calibration standard.

Making Mixer Measurements (Option 089 Only) Phase or Group Delay Measurements
An important characteristic to remember when selecting a calibration mixer is that the delay of the device should be kept as low as possible. To do this, select a mixer with very wide bandwidth (wider bandwidth results in smaller delay). The accuracy of this measurement depends on the quality of the mixer that is being used for calibration and how well this mixer has been characterized. The following measurement must be performed with a broadband calibration mixer that has a known group delay. The following table lists the specifications of two mixers that may be used for calibration:

System INSTRUMENT MODE FREQ OFFS MENU
LO FREQUENCY 1000 M/µ

Power 0 PWR RANGE MAN x1 −10 x1
Preset

Making Mixer Measurements (Option 089 Only)
Phase or Group Delay Measurements
Figure 2-24 Connections for a Group Delay Measurement

System INSTRUMENT MODE FREQ OFFS MENU
System INSTRUMENT MODE FREQ OFFS MENU VIEW MEASURE
Meas INPUT PORTS B/R
Cal CALIBRATE MENU RESPONSE THRU
Format DELAY

Making Mixer Measurements (Option 089 Only) Phase or Group Delay Measurements
10.Replace the "calibration" mixer with the device under test. If measuring group delay, set the delay equal to the "calibration" mixer delay (for example −0.6 ns) by pressing:
11.Scale the data for best vertical resolution.
Figure 2-25 Group Delay Measurement Example

The displayed measurement trace shows the device under test delay, relative to the "calibration" mixer. This measurement is also useful when the device under test has built-in filtering, which requires >30 dB range (the maximum of R input). PORT 1 to PORT 2 range is >100 dB.
Scale Ref · −06 G/n · Scale Ref

Making Mixer Measurements (Option 089 Only)
Amplitude and Phase Tracking
The match between mixers is defined as the absolute difference in amplitude or phase response over a specified frequency range. The tracking between mixers is essentially how well the devices are matched over a specified interval. This interval may be a frequency interval or a temperature interval, or a combination of both.
You can make tracking measurements by ratioing the responses of two mixer conversion loss measurements. Then any difference you view in response is due to the mixers and not the measurement system.
Using the same measurement setup as in "Phase or Group Delay Measurements" on page 2-35, you can determine how well two mixers track each other in terms of amplitude and phase.
In step 7, select .
Display DATA→MEM
The resulting trace should represent the amplitude and phase tracking of the two mixers.
Figure 2-26 Connections for an Amplitude and Phase Tracking Measurement
Format PHASE · Between Two Mixers

Making Mixer Measurements (Option 089 Only) Conversion Compression Using the Frequency Offset Mode
Conversion compression is a measure of the maximum RF input signal level where the mixer provides linear operation. The conversion loss is the ratio of the IF output level to the RF input level. This value remains constant over a specified input power range. When the input power level exceeds a certain maximum, the constant ratio between IF and RF power levels will begin to change. The point at which the ratio has decreased by 1 dB is called the 1 dB compression point. See Figure 2-27.
Figure 2-27 Conversion Loss and Output Power as a Function of Input Power

The following example uses a ratio of mixer output to input power and a marker search function to locate a mixer’s 1 dB compression point.
If the default output power of your analyzer is not high enough to force the mixer under test into compression, then the following procedure may have to be performed with the addition of Option 007 or Option 085. Refer to the "Making High Power Measurements with Option 085 (ES Analyzers Only)" on page 1-59.
SWEEP TYPE MENU POWER SWEEP Return
Level Example · Preset · Sweep Setup

Making Mixer Measurements (Option 089 Only) Conversion Compression Using the Frequency Offset Mode
System INSTRUMENT MODE FREQ OFFS MENU
LO FREQUENCY 800 M/µ
Start −10 x1
Figure 2-28 Connections for the First Portion of Conversion Compression

Meas INPUT PORTS R
DATA/MEM
This removes the loss between the output of the mixer and the input to the receiver, and provides a linear power sweep for use in subsequent measurements.

Making Mixer Measurements (Option 089 Only) Conversion Compression Using the Frequency Offset Mode
Figure 2-29 Connections for the Second Portion of Conversion Compression

LO MENU FREQUENCY:CW 600 M/µ
10.To select the converter type, press:
UP CONVERTER
11.To select a low-side LO measurement configuration, press:
FREQ OFFS ON
In this low-side LO, up converter measurement, the analyzer source frequency is offset lower than the receiver frequency. The analyzer source frequency can be determined from the following equation:
receiver frequency (800 MHz) − LO frequency (600 MHz) = 200 MHz
Measurement · System

Making Mixer Measurements (Option 089 Only) Conversion Compression Using the Frequency Offset Mode
The measurements setup diagram is shown in Figure 2-30.
Figure 2-30 Measurement Setup Diagram Shown on Analyzer Display

12.To view the mixer’s output power as a function of its input power, press:
13.To set up an active marker to search for the 1 dB compression point of the mixer, press:
Marker Search SEARCH:MAX
14.Press:
Marker Search TARGET −1 x1
Scale Ref · Marker MKR ZERO

Making Mixer Measurements (Option 089 Only) Conversion Compression Using the Frequency Offset Mode
The measurement results show the mixer’s 1 dB compression point. By changing the target value, you can easily locate other compression points (for example, 0.5 dB, 3 dB). See Figure 2-31.
15.Read the compressed power on by turning marker ∆ off.
Marker ∆ MODE MENU ∆ MODE OFF
Figure 2-31 Example Swept Power Conversion Compression Measurement



Isolation is the measure of signal leakage in a mixer. Feedthrough is specifically the forward signal leakage to the IF port. High isolation means that the amount of leakage or feedthrough between the mixer’s ports is very small. Isolation measurements do not use the frequency offset mode. Figure 2-32 illustrates the signal flow in a mixer.
Figure 2-32 Signal Flow in a Mixer Example

LO to RF isolation is the amount the LO power is attenuated when it appears directly at the RF port.

Power 0 PWR RANGE MAN x1 −10 x1
This source stimulates the mixer’s LO port.
Meas INPUT PORTS B/R
Preset

Making Mixer Measurements (Option 089 Only) Isolation Example Measurements
Figure 2-33 Connections for a Response Calibration

Cal CALIBRATE MENU RESPONSE THRU
Figure 2-34 Connections for a Mixer Isolation Measurement


Making Mixer Measurements (Option 089 Only)
Scale Ref AUTO SCALE
The measurement results show the mixer’s LO to RF isolation.
Figure 2-35 Example Mixer LO to RF Isolation Measurement

The procedure and equipment configuration necessary for this measurement are very similar to those of the previous LO to RF Isolation procedure, with the addition of an external source to drive the mixer’s LO port as we measure the mixer’s RF feedthrough. RF feedthrough measurements do not use the frequency offset mode.
CW frequency = 300 MHz
Power = 10 dBm

Power 0 PWR RANGE MAN x1 −10 x1
This source stimulates the mixer’s LO port.
Meas INPUT PORTS B/R
Isolation Example Measurements · Preset

Making Mixer Measurements (Option 089 Only) Isolation Example Measurements
Figure 2-36 Connections for a Response Calibration
Cal CALIBRATE MENU RESPONSE THRU
Figure 2-37 Connections for a Mixer RF Feedthrough Measurement


Making Mixer Measurements (Option 089 Only)
Figure 2-38 Example Mixer RF Feedthrough Measurement

You can measure the IF to RF isolation in a similar manner, but with the following modifications:
Isolation Example Measurements

Making Mixer Measurements (Option 089 Only) Isolation Example Measurements
Reflection coefficient (Γ) is defined as the ratio between the reflected voltage (Vr) and incident voltage (Vi). Standing wave ratio (SWR) is defined as the ratio of maximum standing wave voltage to the minimum standing wave voltage and can be derived from the reflection coefficient (Γ) using the following equation. Return loss can be derived from the reflection coefficient as well.
Γ Vr Vi =
SWR 1 Γ + 1 Γ – =
Return loss 20 – Γ log =
When you measure the RF port SWR, you should have the LO drive level present and set to the expected frequency and power levels. Different LO drives and frequencies may yield different values for SWR at the same RF frequencies. The IF port should be terminated in a condition as close to its operating state as possible.
The measurements of LO port SWR and IF port SWR are very similar. For IF port SWR, you should terminate the RF port in a matched condition and apply the LO signal at its normal operating level. For the LO port SWR, the RF and IF ports should both be terminated in conditions similar to what will be present during normal operation.


Making Time Domain Measurements Using This Chapter
This chapter contains the following:
— "Time Domain Bandpass Mode" on page 3-12
— "Time Domain Low Pass Mode" on page 3-15
— "Transforming CW Time Measurements into the Frequency Domain" on page 3-22
— "Masking" on page 3-26
— "Windowing" on page 3-27
— "Range" on page 3-30
— "Resolution" on page 3-32
— "Gating" on page 3-35

Making Time Domain Measurements Introduction to Time Domain Measurements
The analyzers with Option 010 allow you to measure the time domain response of a device. Time domain analysis is useful for isolating a device problem in time or in distance. Time and distance are related by the velocity factor of your device under test (DUT) which is described in "Time Domain Bandpass Mode" on page 3-12.
The analyzer measures the frequency response of your device and uses an inverse Fourier transform (a mathematical calculation) to convert the frequency domain information into the time domain, with time as the horizontal display axis. The analyzer’s internal computer makes this mathematical calculation using the chirp-Z Fourier transform technique.
The resulting measurement is the fully error-corrected time domain reflection or transmission response of the test device, displayed in near real-time. Response values (measured on the vertical axis) now appear separated in time or distance, providing valuable insight into the behavior of the test device beyond simple frequency characteristics.
With Option 010, the analyzer can transform frequency domain data to the time domain or time domain data to the frequency domain.
The transform used by the analyzer resembles time domain reflectometry (TDR) measurements. TDR measurements, however, are made by launching an impulse or step into the test device and observing the response in time with a receiver similar to an oscilloscope. In contrast, the analyzer makes swept frequency response measurements, and mathematically transforms the data into a TDR-like display.
Figure 3-1 illustrates the frequency and time domain reflection responses of a test device. The frequency domain reflection measurement is the composite of all the signals reflected by the discontinuities present in the test device over the measured frequency range.

Making Time Domain Measurements Introduction to Time Domain Measurements
Figure 3-1 Device Frequency Domain and Time Domain Reflection Responses

The time domain measurement shows the effect of each discontinuity as a function of time (or distance), and shows that the test device response consists of three separate impedance changes. The second discontinuity has a reflection coefficient magnitude of 0.035 (i.e. 3.5% of the incident signal is reflected). Marker 1 on the time domain trace shows the elapsed time from the reference plane (where the calibration standards are connected) to the discontinuity and back: 18.2 nanoseconds.
The analyzer has three frequency-to-time transform modes:
The analyzer has one time-to-frequency transform mode:


In this example measurement there are three components of the transmission response:
This example procedure also shows you how time domain analysis allows you to mathematically remove individual parts of the time domain response to see the effect of potential design changes. This is accomplished by "gating" out the undesirable responses. With the "gating" capability, the analyzer time domain allows you to perform "what if" analysis by mathematically removing selected reflections and seeing the effect in the frequency domain.
Figure 3-2 Device Connections for Time Domain Transmission Example Measurement



Making Time Domain Measurements Making Transmission Response Measurements
System TRANSFORM MENU BANDPASS TRANSFORM ON
Start 0 G/n Stop 6 M/µ
The other time domain modes, low pass step and low pass impulse, are described in "Time Domain Low Pass Mode" on page 3-15.
and turn the front panel knob, or enter a value from the front panel keypad.
Scale Ref REFERENCE VALUE
The three responses shown in Figure 3-3 are the RF leakage near zero seconds, the main travel path through the filter, and the triple travel path through the filter. Only the combination of these responses was evident to you in the frequency domain.
Figure 3-3 Time Domain Transmission Example Measurement

System TRANSFORM MENU SPECIFY GATE CENTER
, or turn the front panel knob to position the center gate marker. This marker, shaped like a “T”, is shown in Figure 3-4.
1.6 M/µ
10.To set the gate span, press:
Span 1.2 M/µ
or turn the front panel knob to position the "flag" gate markers.

Making Time Domain Measurements Making Transmission Response Measurements
11.To activate the gating function to remove any unwanted responses, press:
As shown in Figure 3-4, only response from the main path is displayed.
Figure 3-4 Gating in a Time Domain Transmission Example Measurement

12.To adjust the gate shape for the best possible time domain response, press and select between minimum, normal, wide, and maximum. Each gate has a different passband flatness, cutoff rate, and sidelobe levels. A detailed discussion of gating and gate shape selections is located in "Gating" on page 3-35 and "Selecting Gate Shape" on page 3-36.
The passband ripple and sidelobe levels are descriptive of the gate shape. The cutoff time is the time between the stop time (−6 dB on the filter skirt) and the peak of the first sidelobe, and is equal on the left and right side skirts of the filter. The minimum gate span is just twice the cutoff time because it has no passband.

Making Time Domain Measurements Making Transmission Response Measurements
Figure 3-5 Gate Shape
System TRANSFORM MENU TRANSFORM OFF
Scale Ref AUTO SCALE
System TRANSFORM MENU SPECIFY GATE GATE OFF
This places the gated response in memory. Figure 3-6 shows the effect of removing the RF leakage and the triple travel signal path using gating. By transforming back to the frequency domain, we see that this design change would yield better out-of-band rejection.
Figure 3-6 Gating Effects in a Frequency Domain Example Measurement



The time domain response of a reflection measurement is often compared with the time domain reflectometry (TDR) measurements. Like the TDR, the analyzer measures the size of the reflections versus time (or distance). Unlike the TDR, the time domain capability of the analyzer allows you to choose the frequency range over which you would like to make the measurement.
or on ET models:
Start 50 M/µ
Figure 3-7 Device Connections for Reflection Time Domain Example Measurement

Scale Ref AUTO SCALE
Figure 3-8 shows the frequency domain reflection response of the cables under test. The complex ripple pattern is caused by reflections from the adapters interacting with each other. By transforming this data to the time domain, you can determine the magnitude of the reflections versus distance along the cable.
Preset · Stop 3 G/n

Making Time Domain Measurements Making Reflection Response Measurements
Figure 3-8 Device Response in the Frequency Domain

System TRANSFORM MENU BANDPASS TRANSFORM ON
The stop time corresponds to the length of the cable under test. The energy travels about 1 foot per nanosecond, or 0.3 meter/ns, in free space. Most cables have a relative velocity of about 0.66 the speed in free space. Calculate about 3 ns/foot, or 10 ns/meter, for the stop time when you are measuring the return trip distance to the cable end.
Cal MORE VELOCITY FACTOR
NOTE Most cables have a relative velocity of 0.66 (for polyethylene dielectrics) or 0.7 (for teflon dielectrics). If you would like the markers to read actual one-way distance rather than return trip distance, enter one-half the actual velocity factor. Then the markers will read the actual one-way distance to the reflection of interest rather than the "electrical length" that assumes a relative velocity of 1.
Velocity Factor 1 εr =
where εr is the relative permittivity of the cable dielectric.
Format LIN MAG · Start 0 x1 · Stop 35 G/n

Making Time Domain Measurements Making Reflection Response Measurements
and turn the front panel knob, or enter a value from the front panel keypad.
In this example, the velocity factor was set to one-half the actual value, so the marker reads the time and distance to the reflection.
, turning the front panel knob or entering a value from the front panel keypad after each key press.
Figure 3-9 Device Response in the Time Domain

Marker

Making Time Domain Measurements Time Domain Bandpass Mode
This mode is called bandpass because it works with band-limited devices. Traditional TDR requires that the test device be able to operate down to dc. Using bandpass mode, there are no restrictions on the measurement frequency range. Bandpass mode characterizes the test device impulse response.
A marker provides both the two-way time and the two-way electrical length (or distance) to a discontinuity. The distance displayed is based on the assumption that the signal travels at the speed of light. The signal travels slower than the speed of light in most media (e.g. coax cables). This slower velocity (relative to light) can be compensated for by adjusting the analyzer relative velocity factor. To determine the physical length, rather than the electrical length, change the velocity factor to that of the medium under test:
Cal MORE VELOCITY FACTOR
The bandpass mode can transform reflection measurements to the time domain. Figure 3-10 (left) shows a typical frequency response reflection measurement of two sections of cable. Figure 3-10 (right) shows the same two sections of cable in the time domain using the bandpass mode.

Making Time Domain Measurements
Time Domain Bandpass Mode
Figure 3-10 A Reflection Measurement of Two Cables
The ripples in reflection coefficient versus frequency in the frequency domain measurement are caused by the reflections at each connector "beating" against each other.
One at a time, loosen the connectors at each end of the cable and observe the response in both the frequency domain and the time domain. The frequency domain ripples increase as each connector is loosened, corresponding to a larger reflection adding in and out of phase with the other reflections. The time domain responses increase as you loosen the connector that corresponds to each response.
Interpreting the Bandpass Reflection Response Horizontal Axis
In bandpass reflection measurements, the horizontal axis represents the time it takes for an impulse launched at the test port to reach a discontinuity and return to the test port (the two-way travel time). In Figure 3-10, each connector is a discontinuity.
Interpreting the Bandpass Reflection Response Vertical Axis
The quantity displayed on the vertical axis depends on the selected format. The common formats are listed in Table 3-1. The default format is LOG MAG (logarithmic magnitude), which displays the return loss in decibels (dB). LIN MAG (linear magnitude) is a format that displays the response as reflection coefficient ρ. This can be thought of as an average reflection coefficient of the discontinuity over the frequency range of the measurement. Use the REAL format only in low pass mode.

Making Time Domain Measurements Time Domain Bandpass Mode
Table 3-1 Time Domain Reflection Formats
Reflection Coefficient (unitless) (0 < ρ < 1)
Reflection Coefficient (unitless) (−1 < ρ < 1)
LOG MAG
SWR
Standing Wave Ratio (unitless)
The bandpass mode can also transform transmission measurements to the time domain. For example, this mode can provide information about a surface acoustic wave (SAW) filter that is not apparent in the frequency domain. Figure 3-11 illustrates a time domain bandpass measurement of a 321 MHz SAW filter.
Figure 3-11 Transmission Measurement in Time Domain Bandpass Mode
Interpreting the Bandpass Transmission Response Horizontal Axis
In time domain transmission measurements, the horizontal axis is displayed in units of time. The time axis indicates the propagation delay through the device. Note that in time domain transmission measurements, the value displayed is the actual delay (not twice the delay). The marker provides the propagation delay in both time and distance.
Marker 2 in Figure 3-11 (left) indicates the main path response through the test device, which has a propagation delay of 655.6 ns, or about 196.5 meters in electrical length. Marker 4 in Figure 3-11 (right) indicates the triple-travel path response at 1.91 µs, or about 573.5 meters. The response at marker 1 (at 0 seconds) is an RF feedthrough leakage path. In addition to the triple travel path response, there are several other multi-path responses through the test device, which are inherent in the design of a SAW filter.
Interpreting the Bandpass Transmission Response Vertical Axis
In the log magnitude format, the vertical axis displays the transmission loss or gain in dB; in the linear magnitude format it displays the transmission coefficient τ. Think of this as an average of the transmission response over the measurement frequency range.
Format Parameter · Return Loss (dB)

Making Time Domain Measurements
Time Domain Low Pass Mode
This mode is used to simulate a traditional time domain reflectometry (TDR) measurement. It provides information to determine the type of discontinuity (resistive, capacitive, or inductive) that is present. Low pass provides the best resolution for a given bandwidth in the frequency domain. It may be used to give either the step or impulse response of the test device.
The low pass mode is less general-purpose than the bandpass mode because it places strict limitations on the measurement frequency range. The low pass mode requires that the frequency domain data points are harmonically related from dc to the stop frequency. That is, stop = n × start, where n = number of points.
For example, with a start frequency of 50 MHz and 101 points, the stop frequency would be 5.05 GHz. Since the analyzer frequency range starts at 50 MHz, the dc frequency response is extrapolated from the lower frequency data. The requirement to pass dc is the same limitation that exists for traditional TDR.
Before a low pass measurement is made, the measurement frequency range must meet the (stop = n × start) requirement previously described. The softkey performs this function automatically: the stop frequency is set close to the entered stop frequency, and the start frequency is set equal to stop/n.
If the low end of the measurement frequency range is critical, it is best to calculate approximate values for the start and stop frequencies before pressing
and calibrating. This avoids distortion of the measurement results. To see an example, select the preset values of 201 points and a 50 MHz to 13.5 GHz frequency range. Now press and observe the change in frequency values. The stop frequency changes to 13.499 GHz, and the start frequency changes to 67.164 MHz. This would cause a distortion of measurement results for frequencies from 50 MHz to 67.164 MHz.
SET FREQ LOW PASS
NOTE If the start and stop frequencies do not conform to the low pass requirement before a low pass mode (step or impulse) is selected and transform is turned on, the analyzer resets the start and stop frequencies. If error correction is on when the frequency range is changed, this turns it off. Therefore, set the frequency range for time domain low pass before performing a calibration.
Table 3-2 Minimum Frequency Ranges for Time Domain Low Pass
Number of Points Minimum Frequency Range Number of Points Minimum Frequency Range
3 50 MHz to 150 MHz 101 50 MHz to 5.05 GHz
11 50 MHz to 550 MHz 201 50 MHz to 10.05 GHz
21 50 MHz to 1.05 GHz 401 50 MHz to 20.05 GHz

Making Time Domain Measurements Time Domain Low Pass Mode
Table 3-2 Minimum Frequency Ranges for Time Domain Low Pass
Number of Points Minimum Frequency Range Number of Points Minimum Frequency Range
51 50 MHz to 2.55 GHz 801 50 MHz to 40.05 GHz
Minimum Allowable Stop Frequencies
The lowest analyzer measurement frequency is 50 MHz, therefore for each value of n there is a minimum allowable stop frequency that can be used. That is, the minimum stop frequency = n × 50 MHz. Table 3-2 lists the minimum frequency range that can be used for each value of n for low pass time domain measurements.
Figure 3-12 shows the time domain response of an unterminated cable in both the low-pass step and low-pass impulse modes.
Figure 3-12 Time Domain Low Pass Measurements of an Unterminated Cable

Interpreting the Low Pass Response Horizontal Axis
The low pass measurement horizontal axis is the two-way travel time to the discontinuity (as in the bandpass mode). The marker displays both the two-way time and the electrical length along the trace. To determine the actual physical length, enter the appropriate velocity factor as described in "Time Domain Bandpass Mode" on page 3-12.
Interpreting the Low Pass Response Vertical Axis
The vertical axis depends on the chosen format. In the low pass mode, the frequency domain data is taken at harmonically related frequencies and extrapolated to dc. Because this results in the inverse Fourier transform having only a real part (the imaginary part is zero), the most useful low pass step mode format in this application is the real format. It displays the response in reflection coefficient units. This mode is similar to the traditional TDR response, which displays the reflected signal in a real format (volts) versus time (or distance) on the horizontal axis.

Making Time Domain Measurements
Time Domain Low Pass Mode
The real format can also be used in the low pass impulse mode, but for the best dynamic range for simultaneously viewing large and small discontinuities, use the log magnitude format.

Making Time Domain Measurements Time Domain Low Pass Mode
As described, the low pass mode can simulate the TDR response of the test device. This response contains information useful in determining the type of discontinuity present.
Figure 3-13 illustrates the low pass responses of known discontinuities. Each circuit element was simulated to show the corresponding low pass time domain S11 response waveform. The low pass mode gives the test device response either to a step or to an impulse stimulus. Mathematically, the low pass impulse stimulus is the derivative of the step stimulus.
Figure 3-13 Simulated Low Pass Step and Impulse Response Waveforms
Figure 3-14 shows example cables with discontinuities (faults) using the low pass step mode with the real format.
(Real Format)

Making Time Domain Measurements
Time Domain Low Pass Mode
Figure 3-14 Low Pass Step Measurements of Common Cable Faults
Measuring Small Signal Transient Response Using Low Pass Step
Use the low pass mode to analyze the test device’s small signal transient response. The transmission response of a device to a step input is often measured at lower frequencies, using a function generator (to provide the step to the test device) and a sampling oscilloscope (to analyze the test device output response). The low pass step mode extends the frequency range of this type of measurement to the maximum frequency of the analyzer.
The step input shown in Figure 3-15 is the inverse Fourier transform of the frequency domain response of a thru measured at calibration. The step rise time is proportional to the highest frequency in the frequency domain sweep; the higher the frequency, the faster the rise time. The frequency sweep in Figure 3-15 is from 10 MHz to 1 GHz.
Figure 3-15 also illustrates the time domain low pass response of an amplifier under test. The average group delay over the measurement frequency range is the difference in time between the step and the amplifier response. This time domain response simulates an oscilloscope measurement of the amplifier’s small signal transient response. Note the ringing in the amplifier response that indicates an under-damped design.
(Real Format)

Making Time Domain Measurements Time Domain Low Pass Mode
Figure 3-15 Time Domain Low Pass Measurement of an Amplifier Small Signal

Interpreting the Low Pass Step Transmission Response Horizontal Axis
The low pass transmission measurement horizontal axis displays the average transit time through the test device over the frequency range used in the measurement. The response of the thru connection used in the calibration is a step that reaches 50% unit height at approximately time = 0. The rise time is determined by the highest frequency used in the frequency domain measurement. The step is a unit high step, which indicates no loss for the thru calibration. When a device is inserted, the time axis indicates the propagation delay or electrical length of the device. The markers read the electrical delay in both time and distance. The distance can be scaled by an appropriate velocity factor as described in "Time Domain Bandpass Mode" on page 3-12.
Interpreting the Low Pass Step Transmission Response Vertical Axis
In the real format, the vertical axis displays the transmission response in real units (for example, volts). For the amplifier example in Figure 3-15, if the amplifier input is a step of 1 volt, the output, 2.4 nanoseconds after the step (indicated by marker 1), is 5.84 volts.
In the log magnitude format, the amplifier gain is the steady state value displayed after the initial transients die out.
Measuring Separate Transmission Paths through the Test Device Using Low Pass Impulse Mode
The low pass impulse mode can be used to identify different transmission paths through a test device that has a response at frequencies down to dc (or at least has a predictable response, above the noise floor, below 50 MHz).
For example, use the low pass impulse mode to measure the relative transmission times through a multi-path device such as a power divider. Another example is to measure the pulse dispersion through a broadband transmission line, such as a fiber optic cable. Both examples are illustrated in Figure 3-16. The horizontal and vertical axes can be interpreted as already described in "Time Domain Bandpass Mode" on page 3-12.
Transient Response

Making Time Domain Measurements
Time Domain Low Pass Mode
Figure 3-16 Transmission Measurements Using Low Pass Impulse Mode


Making Time Domain Measurements Transforming CW Time Measurements into the Frequency Domain
The analyzer can display the amplitude and phase of CW signals versus time. For example, use this mode for measurements such as amplifier gain as a function of warmup time (i.e. drift). The analyzer can display the measured parameter (e.g. amplifier gain) for periods of up to 24 hours and then output the data to a digital plotter for hardcopy results.
These "strip chart" plots are actually measurements as a function of time (time is the independent variable), and the horizontal display axis is scaled in time units. Transforms of these measurements result in frequency domain data. Such transforms are called forward transforms because the transform from time to frequency is a forward Fourier transform, and can be used to measure the spectral content of a CW signal. For example, when transformed into the frequency domain, a pure CW signal measured over time appears as a single frequency spike. The transform into the frequency domain yields a display that looks similar to a spectrum analyzer display of signal amplitude versus frequency.
Figure 3-17 shows an example of a measurement using the Fourier transform in the forward direction, from the time domain to the frequency domain.
Figure 3-17 Amplifier Gain Measurement

Interpreting the Forward Transform Vertical Axis
With the log magnitude format selected, the vertical axis displays dB. This format simulates a spectrum analyzer display of power versus frequency.

Making Time Domain Measurements Transforming CW Time Measurements into the Frequency Domain
Interpreting the Forward Transform Horizontal Axis
In a frequency domain transform of a CW time measurement, the horizontal axis is measured in units of frequency. The center frequency is the offset of the CW frequency. For example, with a center frequency of 0 Hz, the CW frequency (250 MHz in the example) is in the center of the display. If the center frequency entered is a positive value, the CW frequency shifts to the right half of the display; a negative value shifts it to the left half of the display. The span value entered with the transform on is the total frequency span shown on the display. (Alternatively, the frequency display values can be entered as start and stop.)
Demodulating the Results of the Forward Transform
The forward transform can separate the effects of the CW frequency modulation amplitude and phase components. For example, if a test device modulates the transmission response (S21) with a 500 Hz AM signal, you can see the effects of that modulation as shown in Figure 3-18. To simulate this effect, apply a 500 Hz sine wave to the analyzer rear panel EXT AM input.
Figure 3-18 Combined Effects of Amplitude and Phase Modulation

Using the demodulation capabilities of the analyzer, it is possible to view the amplitude or the phase component of the modulation separately. The window menu includes the following softkeys to control the demodulation feature:
is the normal preset state, in which both the amplitude and phase
components of any test device modulation appear on the display.
displays only the amplitude modulation, as illustrated in Figure 3-19a.
displays only the phase modulation, as shown in Figure 3-19b.

Making Time Domain Measurements Transforming CW Time Measurements into the Frequency Domain
Figure 3-19 Separating the Amplitude and Phase Components of

In the forward transform (from CW time to the frequency domain), range is defined as the frequency span that can be displayed before aliasing occurs, and is similar to range as defined for time domain measurements. In the range formula, substitute time span for frequency span.
Range Number of points 1 –
time span =
Range 201 1 –
200 10 3 – × =
Range 1000 Hertz =
For the example, a 201 point CW time measurement made over a 200 ms time span, choose a span of 1 kHz or less on either side of the center frequency (see Figure 3-20). That is, choose a total span of 2 kHz or less.
Test-Device-Induced Modulation · Forward Transform Range

Making Time Domain Measurements Transforming CW Time Measurements into the Frequency Domain
Figure 3-20 Range of a Forward Transform Measurement

To increase the frequency domain measurement range, increase the span. The maximum range is inversely proportional to the sweep time, therefore it may be necessary to increase the number of points or decrease the sweep time. Because increasing the number of points increases the auto sweep time, the maximum range is 2 kHz on either side of the selected CW time measurement center frequency (4 kHz total span). To display a total frequency span of 4 kHz, enter the span as 4000 Hz.

Making Time Domain Measurements Masking
Masking occurs when a discontinuity (fault) closest to the reference plane affects the response of each subsequent discontinuity. This happens because the energy reflected from the first discontinuity never reaches subsequent discontinuities. For example, if a transmission line has two discontinuities that each reflect 50% of the incident voltage, the time domain response (real format) shows the correct reflection coefficient for the first discontinuity (ρ=0.50). However, the second discontinuity appears as a 37.5% reflection (ρ=0.375) because only some the incident voltage reached the second discontinuity, and some of that reflected energy is reflected off the first discontinuity as it returns. For two discrete discontinuities, the apparent reflection of the second discontinuity is appears as
2 ⋅ = ρˆ
approximately , where is the apparent reflection of the second
discontinuity, is the reflection of the first discontinuity, and is the reflection of the second discontinuity.
ρ1 ρ2
As an example of masking due to line loss, consider the time domain response of a 3 dB attenuator and a short circuit. The impulse response (log magnitude format) of the short circuit alone is a return loss of 0 dB, as shown in Figure 3-21a. When the short circuit is placed at the end of the 3 dB attenuator, the return loss is −6 dB, as shown in Figure 3-21b. This value actually represents the forward and return path loss through the attenuator, and illustrates how a lossy network can affect the responses that follow it.
Figure 3-21 Masking Example

ρˆ · 2 1 ρ

Making Time Domain Measurements
The analyzer provides a windowing feature that makes time domain measurements more useful for isolating and identifying individual responses. Windowing is needed because of the abrupt transitions in a frequency domain measurement at the start and stop frequencies. The band limiting of a frequency domain response causes overshoot and ringing in the time domain response, and causes a non-windowed impulse stimulus to have a sin(kt)/kt shape, where k = π/frequency span and t = time (see Figure 3-22). This has two effects that limit the usefulness of the time domain measurement:
Figure 3-22 Impulse Width, Sidelobes, and Windowing
Windowing improves the dynamic range of a time domain measurement by filtering the frequency domain data prior to converting it to the time domain, producing an impulse stimulus that has lower sidelobes. This makes it much easier to see time domain responses that are very different in magnitude. The sidelobe reduction is achieved, however, at the expense of increased impulse width. The effect of windowing on the step stimulus (low pass mode only) is a reduction of overshoot and ringing at the expense of increased rise time.
To select a window, press . A menu is presented that allows the selection of three window types, see Table 3-3.
System TRANSFORM MENU WINDOW
Windowing

Making Time Domain Measurements Windowing
Table 3-3 Impulse Width, Sidelobe Level, and Windowing Values
Window Type Impulse Sidelobe
Minimum −13 dB 0.60/Freq Span −21 dB 0.45/Freq Span
Normal −44 dB 0.98/Freq Span −60 dB 0.99/Freq Span
Maximum −75 dB 1.39/Freq Span −70 dB 1.48/Freq Span
Choose one of the three window shapes listed or use the knob to select any windowing pulse width (or rise time for a step stimulus) between the softkey values. The time domain stimulus sidelobe levels depend only on the window selected.
is essentially no window. Consequently, it gives the highest sidelobes.
(the preset mode) gives reduced sidelobes and is the mode most often used.
window gives the minimum sidelobes, providing the greatest dynamic range.
USE MEMORY on OFF
remembers a user-specified window pulse width (or step rise time) different from the standard window values.
A window is activated only for viewing a time domain response, and does not affect a displayed frequency domain response. Figure 3-23 shows the typical effects of windowing on the time domain response of a short circuit reflection measurement.

Making Time Domain Measurements
Figure 3-23 The Effects of Windowing on the Time Domain Responses of a Short
Windowing · Circuit (Real Format)

Making Time Domain Measurements Range
In the time domain, range is defined as the length in time that a measurement can be made without encountering a repetition of the response, called aliasing. A time domain response repeats at regular intervals because the frequency domain data is taken at discrete frequency points, rather than continuously over the frequency band.
Measurement range 1 ∆F =
where ∆F is the spacing between frequency data points
number of points 1 – ( )
frequency span Hz ( ) =
For example:
Measurement 201 points =
Frequency Span = 1 GHz to 3 GHz
Frequency Spacing ∆F ( ) 10 MHz =
Range 1 ∆F or number of points 1 – ( )
frequency span =
10 106 × ( ) = or 201 1 – ( )
2 109 × ( )
Range 100 10 9 – × seconds =
Electrical Length range the speed of light 3 108m s--- × × =
ElectricalLength 100 10 9 – s × 3 108 m s--- × × =
ElectricalLength 30 meters =
Measurement range · Range 1

Making Time Domain Measurements
In this example, the range is 100 ns, or 30 meters electrical length. To prevent the time domain responses from overlapping, the test device must be 30 meters or less in electrical length for a transmission measurement (15 meters for a reflection measurement). The analyzer limits the stop time to prevent the display of aliased responses.
To increase the time domain measurement range, first increase the number of points, but remember that as the number of points increases, the sweep speed decreases. Decreasing the frequency span also increases range, but reduces resolution.
Range

Making Time Domain Measurements Resolution
Two different resolution terms are used in the time domain:
Time domain response resolution is defined as the ability to resolve two closely-spaced responses, or a measure of how close two responses can be to each other and still be distinguished from each other. For responses of equal amplitude, the response resolution is equal to the 50% (−6 dB) impulse width. It is inversely proportional to the measurement frequency span, and is also a function of the window used in the transform. The approximate formulas for calculating the 50% impulse width are given in Table 3-3. For example, using the formula for the bandpass mode with a normal windowing function for a 50 MHz to 13.05 GHz measurement (13.0 GHz span):
50 percent calculated impulse width 0.98 13.0 GHz ( ) 2 × =
0.151 nanoseconds =
Electrical length in air ( ) 0.151 10 9 – s × 30 109 cm s--- × × =
4.53 centimeters =
With this measurement, two equal responses can be distinguished when they are separated by at least 4.53 centimeters. In a measurement with a 20 GHz span, two equal responses can be distinguished when they are separated by at least 2.94 cm. Using the low pass mode (the low pass frequencies are slightly different) with a minimum windowing function, you can distinguish two equal responses that are about 1.38 centimeters or more apart.
For reflection measurements, which measure the two-way time to the response, divide the response resolution by 2. Using this example, you can distinguish two faults of equal magnitude provided they are 0.69 centimeters (electrical length) or more apart.

Making Time Domain Measurements
For example, a cable with a teflon dielectric (0.7 relative velocity factor), measured under the conditions stated above, has a fault location measurement response resolution of 0.45 centimeters. This is the maximum fault location response resolution. Factors such as reduced frequency span, greater frequency domain data windowing, and a large discontinuity shadowing the response of a smaller discontinuity, all act to degrade the effective response resolution.
Figure 3-24 illustrates the effects of response resolution. The solid line shows the actual reflection measurement of two approximately equal discontinuities (the input and output of an SMA barrel). The dashed line shows the approximate effect of each discontinuity, if they could be measured separately.
Figure 3-24 Response Resolution
While increasing the frequency span increases the response resolution, keep the following points in mind:
You may (with these limitations in mind) choose to use a frequency span that is wider than the test device bandwidth to achieve better resolution.
Resolution

Making Time Domain Measurements Resolution
Time domain range resolution is defined as the ability to locate a single response in time. If only one response is present, range resolution is a measure of how closely you can pinpoint the peak of that response. The range resolution is equal to the digital resolution of the display, which is the time domain span divided by the number of points on the display. To get the maximum range resolution, center the response on the display and reduce the time domain span. The range resolution is always much finer than the response resolution (see Figure 3-25).
Figure 3-25 Range Resolution of a Single Discontinuity

Making Time Domain Measurements
Gating provides the flexibility of selectively removing time domain responses. The remaining time domain responses can then be transformed back to the frequency domain. For reflection (or fault location) measurements, use this feature to remove the effects of unwanted discontinuities in the time domain. You can then view the frequency response of the remaining discontinuities. In a transmission measurement, you can remove the effects of multiple transmission paths.
Figure 3-26a shows the frequency response of an electrical airline and termination. Figure 3-26b shows the response in the time domain. The discontinuity on the left is due to the input connector. The discontinuity on the right is due to the termination. We want to remove the effect of the connector so that we can see the frequency response of just the airline and termination. Figure 3-26c shows the gate applied to the connector discontinuity. Figure 3-26d shows the frequency response of the airline and termination, with the connector "gated out."
Figure 3-26 Sequence of Steps in Gating Operation
Think of a gate as a bandpass filter in the time domain (see Figure 3-27). When the gate is on, responses outside the gate are mathematically removed from the time domain trace. Enter the gate position as a start and stop time (not frequency) or as a center and span time. The start and stop times are the bandpass filter −6 dB cutoff times. Gates can have a negative span, in which case the responses inside the gate are mathematically removed. The gate’s start and stop flags define the region where gating is on.
Gating

Making Time Domain Measurements Gating
Figure 3-27 Gate Shape
The four gate shapes available are listed in Table 3-4. Each gate has a different passband flatness, cutoff rate, and sidelobe levels.
Table 3-4 Gate Characteristics
Gate Shape Passband Ripple Sidelobe Levels Cutoff Time Minimum Gate Span
Minimum ±0.10 dB −48 dB 1.4/Freq Span 2.8/Freq Span
Normal ±0.10 dB −68 dB 2.8/Freq Span 5.6/Freq Span
Wide ±0.10 dB −57 dB 4.4/Freq Span 8.8/Freq Span
Maximum ±0.10 dB −70 dB 12.7/Freq Span 25.4/Freq Span
The passband ripple and sidelobe levels are descriptive of the gate shape. The cutoff time is the time between the stop time (−6 dB on the filter skirt) and the peak of the first sidelobe, and is equal on the left and right side skirts of the filter. As shown in Table 3-4, the minimum gate span is just twice the cutoff time because it has no passband. Always choose a gate span wider than the minimum. For most applications, do not be concerned about the minimum gate span, simply use the knob to position the gate markers around the desired portion of the time domain trace.
Gate Span


Printing, Plotting, and Saving Measurement Results Using This Chapter
This chapter contains instructions for the following tasks:
J Configuring a print function
J Defining a print function
J Printing one measurement per page
J Printing multiple measurements per page
J Configuring a plot function
J Defining a plot function
J Plotting one measurement per page using a pen plotter
J Plotting multiple measurements per page using a pen plotter
J Plotting a measurement to disk
J To view plot files on a PC
J Outputting plot files from a PC to a plotter
J Outputting plot files from a PC to an HP-GL compatible printer
J Outputting single page plots using a printer
J Outputting multiple plots to a single page using a printer
J Plotting multiple measurements per page from disk
J Titling the displayed measurement
J Configuring the analyzer to produce a time stamp
J Aborting a print or plot process
J Printing or plotting the list values or operating parameters
J Solving problems with printing or plotting
J Saving an instrument state
J Saving measurement results
J Re-saving an instrument state
J Deleting a file
J Renaming a file
J Recalling a file
J Formatting a disk
J Solving problems with saving or recalling files

Printing, Plotting, and Saving Measurement Results
Printing or Plotting Your Measurement Results
You can print your measurement results to the following peripherals:
You can plot your measurement results to the following peripherals:
Most Hewlett-Packard desktop printers and plotters are compatible with the analyzer. For a list of recommended peripherals, refer to the configuration guide for your analyzer. The following Web site also contains a link to the configuration guide:
A printer compatibility guide (an up-to-date list of printers that are compatible with the network analyzer) can be found at the following Web site:

Printing, Plotting, and Saving Measurement Results Configuring a Print Function
All copy configuration settings are stored in non-volatile memory. Therefore, they are not affected if you press or switch off the analyzer power.
Figure 4-1 Printer Connections to the Analyzer

Local SET ADDRESSES PRINTER PORT PRNTR TYPE
J DeskJet (This supports most current models such as DeskJet 890C, DeskJet 895C, or
DeskJet 1600C. See also DJ 540 selection.)
J LaserJet (only LaserJet models III, 4, 5, and 6)
J Epson-P2 (printers that conform to the ESC/P2 printer control language, such as
J DJ 540 (This can be used for printers that do not support 100 dots per inch (dpi) but
do support 300 dpi such as HP DeskJet 540 or 850C.)
Information regarding a printer compatibility guide (an up-to-date list of printers that are compatible with the network analyzer) is available in "Printing or Plotting Your Measurement Results" on page 4-3.

Printing, Plotting, and Saving Measurement Results
Configuring a Print Function
a. Enter the GPIB address of the printer, followed by .
b. Press and if there is no external controller connected to the GPIB bus.
c. Press and if there is an external controller connected to the GPIB bus.
Local USE PASS CONTROL
Press and then select the parallel port interface function by pressing until the correct function appears.
J If you choose , the parallel port is dedicated for normal copy device use (printers or plotters).
J If you choose , the parallel port is dedicated for general purpose I/O, and cannot be used for printing or plotting.
a. Press and enter the printer's baud rate, followed by .
b. To select the transmission control method that is compatible with your printer,
press (transmit control - handshaking protocol) until the correct method appears.
J If you choose , the handshake method allows the printer to control the data exchange by transmitting control characters to the network analyzer.
J If you choose , the handshake method allows the printer to control the data exchange by setting the electrical voltage on one line of the RS-232 serial cable.
Local SYSTEM CONTROLLER · Local PARALLEL · Xon-Xoff

Printing, Plotting, and Saving Measurement Results Defining a Print Function
J Choose if you are using a black and white printer, or you want just black and white from a color printer.
J Choose if you are using a color printer.
J Choose if you want to print one measurement per page.
J Choose if you want to print multiple measurements per page.
Since the media color is usually white or clear, you could set a print element to white if you do not want that element to appear on your hardcopy.
Copy DEFINE PRINT

Printing, Plotting, and Saving Measurement Results
Defining a Print Function
Copy DEFINE PRINT DEFAULT PRNT SETUP
Table 4-1 Default Values for Printing Parameters
Channel 1 and 3 Data Magenta
Channel 1 and 3 Memory Green
Channel 2 and 4 Data Blue
Channel 2 and 4 Memory Red

Printing, Plotting, and Saving Measurement Results Printing One Measurement Per Page
If you defined the , press after the message COPY OUTPUT COMPLETED appears.
Copy PRINT MONOCHROME

Printing, Plotting, and Saving Measurement Results
Printing Multiple Measurements Per Page
Copy DEFINE PRINT AUTO-FEED AUTO-FEED OFF
Figure 4-2 Printing Two Measurements
Copy PRINT MONOCHROME

Printing, Plotting, and Saving Measurement Results Configuring a Plot Function
All copy configuration settings are stored in non-volatile memory. Therefore, they are not affected if you press or switch off the analyzer power.
Peripheral Interface Recommended Cables
Figure 4-3 Peripheral Connections to the Analyzer
Local SET ADDRESSES PLTR PORT PLTR TYPE [ ]
Preset · Parallel 92284A · Serial 24542G

Printing, Plotting, and Saving Measurement Results
Configuring a Plot Function
Information regarding a printer compatibility guide (an up-to-date list of printers that are compatible with the network analyzer) is available in "Printing or Plotting Your Measurement Results" on page 4-3.
a. Enter the GPIB address of the printer (default is 01), followed by .
b. Press and if there is no external controller connected to the GPIB bus.
Local USE PASS CONTROL
c. Press and if there is an external controller connected to the GPIB bus.
Press and then select the parallel port interface function by pressing until the correct function appears.
J If you choose , the parallel port is dedicated for normal copy device use (printers or plotters).
J If you choose , the parallel port is dedicated for general purpose I/O, and cannot be used for printing or plotting.
a. Press and enter the printer's baud rate, followed by .
b. To select the transmission control method that is compatible with your printer,
press (transmit control - handshaking protocol) until the correct method appears.
J If you choose , the handshake method allows the printer to control the data exchange by transmitting control characters to the network analyzer.
J If you choose , the handshake method allows the printer to control the data exchange by setting the electrical voltage on one line of the RS-232 serial cable.
Local SET ADDRESSES PLOTTER PORT PLTR TYPE PLTR TYPE [HPGL PRT]
Local SYSTEM CONTROLLER · Local PARALLEL · Xon-Xoff

Printing, Plotting, and Saving Measurement Results Configuring a Plot Function
Local SET ADDRESSES PLOTTER PORT PLTR TYPE PLTR TYPE [PLOTTER]
a. Enter the GPIB address of the plotter (default is 05), followed by .
b. Press and if there is no external controller connected to the GPIB bus.
c. Press and if there is an external controller connected to the GPIB bus.
Local USE PASS CONTROL
J Press and then select the parallel port interface function by pressing until the correct function appears.
— If you choose , the parallel port is dedicated for normal copy device use (printers or plotters).
— If you choose , the parallel port is dedicated for general purpose I/O, and cannot be used for printing or plotting.
a. Press and enter the plotter's baud rate, followed by .
b. To select the transmission control method that is compatible with your plotter,
press (transmit control - handshaking protocol) until the correct method appears.
J If you choose , the handshake method allows the plotter to control the data exchange by transmitting control characters to the network analyzer.
J If you choose , the handshake method allows the plotter to control the data exchange by setting the electrical voltage on one line of the RS-232 serial cable.
Local SYSTEM CONTROLLER · Local PARALLEL · Xon-Xoff

Printing, Plotting, and Saving Measurement Results
Configuring a Plot Function
The plot files that you generate from the analyzer, contain the HPGL representation of the measurement display. The files will not contain any setup or formfeed commands.
a. Press .
Local SET ADDRESSES PLOTTER PORT DISK
b. Press and select the disk drive that you will plot to.
Local DISK UNIT NUMBER
The analyzer assigns the first available default filename for the displayed directory. For example, the analyzer would assign PLOT00FP for a LIF format (PLOT00.FP for a DOS format) if there were no previous plot files saved to the disk.
Figure 4-4 shows the three parts of the file name that are generated automatically by the analyzer whenever a plot is requested. The two digit sequence number is incremented by one each time a file with a default name is added to the directory.
Save/Recall SELECT DISK · Copy PLOT

Printing, Plotting, and Saving Measurement Results Configuring a Plot Function
Figure 4-4 Automatic File Naming Convention for LIF Format
To Output the Plot Files

Printing, Plotting, and Saving Measurement Results
Defining a Plot Function
J Choose if you want the measurement data trace to appear on your plot.
J Choose if you want the displayed memory trace to appear on your plot.
J Choose if you want the graticule and the reference line to appear on your plot.
J Choose if you want all of the displayed text to appear on your plot. (This does not include the marker values or softkey labels.)
J Choose if you want the displayed markers, and marker values, to appear on your plot.
Figure 4-5 Plot Components Available through Definition

J Choose if you want a “page eject” sent to the plotter or HPGL compatible printer after each time you press .
PLOT
J Choose if you want multiple plots on the same sheet of paper.
Copy DEFINE PLOT

Printing, Plotting, and Saving Measurement Results Defining a Plot Function
Selecting Pen Numbers and Colors
PEN NUM DATA
Press after each modification.
Table 4-2 Default Pen Numbers and Corresponding Colors
Pen Number Color Pen Number Color
0 white 4 yellow
1 cyan 5 green
2 magenta 6 red
3 blue 7 black
Table 4-3 Default Pen Numbers for Plot Elements
Corresponding Key Plot Element Pen Numbers
Channel 1 Channel 2
Measurement Data Trace 2 3
Displayed Memory Trace 5 6
Graticule and Reference Line 1 1
PEN NUM TEXT
Displayed Text 7 7
Displayed Markers and Values 7 7
You must define the pen numbers separately for each measurement channel (channel 1/3 and channel 2/4).
![Page 237 of the Agilent Technologies Network Analyzer (20ES) user manual — SCALE PLOT [GRAT]](/cdn/reader/15/150a289e983abebdc4bd6f0c3204b03f577bfd829d08c0cf5bb1353a14e103d7/images/page-237.webp?w=400)
Printing, Plotting, and Saving Measurement Results
Defining a Plot Function
— Select to modify the line type for the data trace. Then enter the new line type (see Figure 4-6), followed by .
— Select to modify the line type for the memory trace. Then enter the new line type (see Figure 4-6), followed by .
Table 4-4 Default Line Types for Plot Elements
Plot Elements Channel 1 Line Type Numbers Channel 2 Line Type Numbers
Data Trace 7 7
Memory Trace 7 7
Figure 4-6 Line Types Available
J Choose if you want the normal scale selection for plotting. This includes space for all display annotations such as marker values and stimulus values. The entire analyzer display fits within the defined boundaries of P1 and P2 on the plotter, while maintaining the exact same aspect ratio as the display.
J Choose if you want the outer limits of the graticule to correspond to the defined P1 and P2 scaling point on the plotter. (Intended for plotting on preprinted rectangular or polar forms.)
![Page 238 of the Agilent Technologies Network Analyzer (20ES) user manual — PLOT SPEED [SLOW]](/cdn/reader/15/150a289e983abebdc4bd6f0c3204b03f577bfd829d08c0cf5bb1353a14e103d7/images/page-238.webp?w=400)
Printing, Plotting, and Saving Measurement Results Defining a Plot Function
Figure 4-7 Locations of P1 and P2 in Mode
Choosing Plot Speed
J Choose for normal plotting.
J Choose for plotting directly on transparencies. (The slower speed provides a more consistent line width.)
To Reset the Plotting Parameters to Default Values
Press .
Copy DEFINE PLOT MORE MORE YES
Table 4-5 Plotting Parameter Default Values
Plotting Parameter Default Value Plotting Parameter Default Value
Select Quadrant Full page Plot Scale Full
Auto Feed ON Plot Speed Fast
Define Plot All plot elements on Line Type 7 (solid line)
Data 2 Data 3
Memory 5 Memory 6
Graticule 1 Graticule 1
Text 7 Text 7
Marker 7 Marker 7

Printing, Plotting, and Saving Measurement Results Plotting One Measurement Per Page Using a Pen Plotter
J If you defined the , press after the message COPY OUTPUT COMPLETED appears.
Copy PLOT

Printing, Plotting, and Saving Measurement Results Plotting Multiple Measurements Per Page Using a Pen Plotter
Copy SEL QUAD [ ]
LEFT UPPER
J
LEFT LOWER
J
RIGHT UPPER
J
RIGHT LOWER
The selected quadrant will appear in the brackets under .
Figure 4-8 Plot Quadrants
Copy

Printing, Plotting, and Saving Measurement Results Plotting Multiple Measurements Per Page Using a Pen Plotter
Copy PLOT PLOTTER FORM FEED

Copy SEL QUAD [ ] SELECT DISK LEFT UPPER PLOT
Seq DO SEQUENCE SEQUENCE 1 SEQ1

Printing, Plotting, and Saving Measurement Results To View Plot Files on a PC
Plot files can be viewed and manipulated on a PC using a word processor or graphics presentation program. Plot files contain a text stream of HPGL (Hewlett-Packard Graphics Language) commands. In order to import a plot file into an application, that application must have an import filter for HPGL (often called HGL). Two such applications from the Lotus® suite of products are the word processor “Ami Pro” and the graphics presentation package “Freelance Graphics.” Additionally, a utility is available to convert plot files to PCX format so they can be used in additional PC applications.
When viewed in such programs, the color and font size of the plot may vary from the output of an HPGL/2 compatible color printer. The following table shows the differences between the color assignments of HPGL/2 compatible printers and Lotus applications. Also refer to "Selecting Pen Numbers and Colors" on page 4-16.
Table 4-6 Color Assignment Differences between HPGL/2-Compatible Printers
HPGL/2 Printer Lotus Applications
Pen Number Color Pen Number Color
0 white N/A N/A
1 cyan (aqua) 1 black
2 magenta (red-violet) 2 red
3 blue 3 green
4 yellow 4 yellow
5 green 5 blue
6 red 6 red-violet (magenta)
7 black 7 aqua (cyan)
To modify the color or font size, consult the documentation for the particular application being used.
and Lotus Applications

Printing, Plotting, and Saving Measurement Results
To View Plot Files on a PC
To view plot files in Ami Pro, perform the following steps:
To view plot files in Freelance, perform the following steps:
a. From the TEXT pull-down menu, select FONT. b. Select the type face and size. (Fourteen point text is a good place to start.) c. Click OK to resize the font.
To change the font color, just do it immediately after you resize the font using the same dialog box.

Printing, Plotting, and Saving Measurement Results Outputting Plot Files from a PC to a Plotter
A utility can convert hpgl (or .fp) files to other PC applications. This utility, named hp2xx, is available to be downloaded without charge (on donation basis only) from Free Software Foundation. You may download this file using the information available on the following Web site: ftp://ots.external.hp.com/rfmw/lou/
To convert HPGL files to be used with other PC applications.
The batch file consists of the following two lines:
@ echo off A:\hp2xx.exe -m pcx %1
(where “A” is the disk drive where the floppy disk in installed.)
Local SET ADDRESSES PLOTTER PORT DISK Copy PLOT

Printing, Plotting, and Saving Measurement Results Outputting Plot Files from a PC to an HPGL Compatible Printer
To output the plot files to an HPGL compatible printer, you can use the HPGL initialization sequence linked in a series as follows:
For example, in the MS-DOS editor (DOS 5.0 or greater), press CNTRL-P (hold down the CTRL key and press P) followed by the ESCape key to create the escape character.
Table 4-7 HPGL Initialization Commands
E conditional page eject
&12A page size 8.5 x 11
&l1O landscape orientation (lower case l, one, capital O)
&a0L no left margin (a, zero, capital L)
&a400M no right margin (a, 4, zero, zero, capital M)
&l0E no top margin (lower case l, zero, capital E)
*c7680x5650Y frame size 10.66"x 7.847" (720 decipoints/inch)
*p50x50Y move cursor to anchor point
*c0T set picture frame anchor point
*r-3U set CMY palette
%1B enter HPGL mode; cursor at PCL
Command Remark

Printing, Plotting, and Saving Measurement Results Outputting Single Page Plots Using a Printer
Table 4-8 HPGL Test File Commands
%0A exit HPGL mode
Type print hpglinit to send the initialization sequence to the printer.
Type print filename (where filename is the name of the HPGL plot file) to send the plot file to the printer.
Type print exithpgl to send the HPGL mode and form feed sequence to the printer.
You can output plot files to an HPGL compatible printer using the DOS command line and the files created in the previous steps. This example assumes that the escape sequence files and the plot files are in the current directory and the selected printer port is PRN.
C:> type hpglinit > PRN
C:> type PLOT00.FP> PRN
C:> type exithpgl > PRN
Command Remark · E form feed · Command Remarks

Printing, Plotting, and Saving Measurement Results Outputting Multiple Plots to a Single Page Using a Printer
Refer to "Plotting Multiple Measurements Per Page Using a Pen Plotter" on page 4-20 for the naming conventions for plot files that you want printed on the same page. You can use the following batch file to automate the plot file printing. In this example, the batch file is be saved as “do_plot.bat.” However, before running this batch file, you must first create the hpglinit file and the exithpgl file described in "Outputting Plot Files from a PC to an HPGL Compatible Printer" on page 4-25.
rem rem Name: do_plot rem rem Description: rem rem output HPGL initialization sequence to a file:spooler rem append all the requested plot files to the spooler rem append the formfeed sequence to the spooler rem copy the file to the printer rem rem (This routine uses COPY instead of PRINT because COPY rem will not return until the action is completed. PRINT rem will queue the file so the subsequent DEL will likely rem generate an error. COPY avoids this.) rem
echo off type hpglinit > spooler for %%i in (%1) do type %%i >> spooler type exithpgl >> spooler copy spooler LPT1 del spooler echo on
For example, you have the following list of files to plot:
PLOT00.LU
PLOT00.RL
PLOT00.RU
You would invoke the batch print as follows:
C:> do_plot PLOT00.*

Printing, Plotting, and Saving Measurement Results Plotting Multiple Measurements Per Page from Disk
The following procedures show you how to store plot files on a LIF formatted disk. A naming convention is used so you can later run an HP BASIC program on an external controller that will output the files to the following peripherals:
The program is provided on the CD-ROM of example programs that is included in the programmer’s guide. The file naming convention allows the program to initiate the following:
The plot file name is made up of four parts; the first three are generated automatically by the analyzer whenever a plot is requested. The two digit sequence number is incremented by one each time a file with a default name is added to the directory.
Figure 4-9 Plot Filename Convention

Printing, Plotting, and Saving Measurement Results Plotting Multiple Measurements Per Page from Disk
You may want to plot various files to the same page, for example, to show measurement data traces for different input settings, or parameters, on the same graticule.
Copy PLOT DEFINE PRINT
For example, you may want only the data trace to appear on the second plot for measurement comparison. In this case, you would press and choose .
Copy DEFINE PLOT PLOT DATA ON PLOT MEM OFF PLOT GRAT OFF PLOT TEXT OFF PLOT MKR OFF
10.Press .
11.Continue defining plots and renaming the saved file until you have saved all the data that you want to put on the same page. Renaming the files as shown allows you to use the provided program, that organizes and plots the files, according to the file naming convention.
Plot File Recognized Filename
First File Saved PLOT00FPA
Second File Saved PLOT00FPB
Third File Saved PLOT00FPC
Fourth File Saved PLOT00FPD
Save/Recall · Copy PLOT · Save/Recall
![Page 250 of the Agilent Technologies Network Analyzer (20ES) user manual — SEL QUAD [ ]](/cdn/reader/15/150a289e983abebdc4bd6f0c3204b03f577bfd829d08c0cf5bb1353a14e103d7/images/page-250.webp?w=400)
Printing, Plotting, and Saving Measurement Results Plotting Multiple Measurements Per Page from Disk
Figure 4-10 shows plots for both the frequency and time domain responses of the same device.
Figure 4-10 Plotting Two Files on the Same Page
Copy SEL QUAD [ ]
Figure 4-11 Plot Quadrants

Printing, Plotting, and Saving Measurement Results Plotting Multiple Measurements Per Page from Disk
Save/Recall

Printing, Plotting, and Saving Measurement Results Titling the Displayed Measurement
a. Turn the front panel knob to move the arrow pointer to the first character of the
title.
b. Press .
c. Repeat the previous two steps to enter the rest of the characters in your title. You
can enter a title that has a maximum of 50 characters.
d. Press to complete the title entry.
Figure 4-12 Example of a Display Title
Display MORE TITLE

Printing, Plotting, and Saving Measurement Results Configuring the Analyzer to Produce a Time Stamp
You can set a clock, and then activate it, if you want the time and date to appear on your hardcopies.
Aborting a Print or Plot Process

Printing, Plotting, and Saving Measurement Results Printing or Plotting the List Values or Operating Parameters
Press and select the information that you want to appear on your hardcopy.
OP PARMS (MKRS etc)
If you are printing the list of measurement data points, each page contains 30 lines of data. The number of pages is determined by the number of measurement points that you have selected.
Choose to print all pages of the listed values.
Copy LIST

Printing, Plotting, and Saving Measurement Results
Solving Problems with Printing or Plotting
If you encounter a problem when you are printing or plotting, check the following list for possible causes:
J connected an interface cable between the peripheral and the analyzer
J connected the peripheral to ac power
J switched on the power
J switched the peripheral on line
J selected the correct printer or plotter type
Local SYSTEM CONTROLLER

Printing, Plotting, and Saving Measurement Results Saving and Recalling Instrument States
The number of registers that the analyzer allows you to save depends on the size of associated error-correction sets, and memory traces. However, the maximum number of registers that can be saved to internal memory is 31. Refer to the "Preset State and Memory Allocation" chapter of the reference guide for further information.
You can save instrument states in the analyzer internal memory, along with the following list of analyzer settings. The default file names are REG.
J number of points

Printing, Plotting, and Saving Measurement Results
Saving and Recalling Instrument States
You can save an instrument state and measurement results to a disk. The default file names are FILEn, where n gets incremented by one each time a file with a default name is added to the directory. The default file names for data-only files are DATAyDz (DATAy.Dz for DOS), where y is incremented by one each time a file with a default name is added to the directory. The z is the channel where the measurement was made. When you save a file to disk, you can choose to save some or all of the following:
Instrument states can be saved to and recalled from an external computer (system controller) using GPIB mnemonics. For more information about the specific analyzer settings that can be saved, refer to the output commands located in the "Command Reference" chapter of the programmer's guide. For an example program, refer to the "Programming Examples" chapter of the programmer's guide.

Printing, Plotting, and Saving Measurement Results Saving an Instrument State
INTERNAL MEMORY
J
INTERNAL DISK
J connect an external disk drive to the analyzer’s GPIB connector, and configure as follows:
a. Connect an external disk drive to the analyzer's GPIB connector, and configure as
follows:
Local DISK UNIT NUMBER
b. Press and enter the drive where your disk is located, followed by .
c. If your storage disk is partitioned, press and enter the volume number where you want to store the instrument state file.
d. Press .
e. Enter the GPIB address of the peripheral, if the default address is incorrect
(default = 00). Follow the entry by pressing .
f. Press and select one of the following:
— allows the analyzer to control peripherals directly.
— allows the computer controller to be involved in all peripheral access operations.
— allows you to control the analyzer over GPIB and also allows the analyzer to take or pass control.
The analyzer saves the state in the next available register, if you are saving to internal memory, or saves the state to disk. Although one file is shown to represent an instrument state on the analyzer display, each instrument state is composed of numerous files (which can be viewed on a PC).
Save/Recall SELECT DISK · Local · Save/Recall SAVE STATE

Printing, Plotting, and Saving Measurement Results
Instrument states combined with measurements results can only be saved to disk. Files that contain data-only, and the various save options available under the
key, are also only valid for disk saves.
The analyzer stores data in arrays along the processing flow of numerical data, from IF detection to display. These arrays are points in the flow path where data is accessible, usually via GPIB. You can choose from three different arrays which vary in modification flexibility when they are recalled.
If you choose to save the raw data array, you will have the most flexibility in modifying the recalled measurement (including the ability to view all four S-parameters). This is because the raw data array has the least amount of processing associated with it. Conversely, if you choose to save the format array, your modification of the recalled measurement will be limited by all the processes that are associated with that measurement result. However, the format array is appropriate if you want to retrieve data traces that look like the currently displayed data.
Define Save Modification Flexibility During Recall
You can also save data-only. This is saved to disk with default file names
DATA00D1 to DATA31D1 for channel 1 DATA00D2 to DATA31D2 for channel 2 DATA00D3 to DATA31D3 for channel 3 DATA00D4 to DATA31D4 for channel 4
However, these files are not instrument states and cannot be recalled.

Printing, Plotting, and Saving Measurement Results Saving Measurement Results
Figure 4-13 Data Processing Flow Diagram
DATA ARRAY ON
J
RAW ARRAY ON
J
FORMAT ARY ON
Save/Recall SELECT DISK · • · Save/Recall DEFINE DISK-SAVE

Printing, Plotting, and Saving Measurement Results
If you select , , or , the data is stored to disk in IEEE-64 bit real format (for LIF disks), and 32 bit PC format for DOS disks. This makes the DOS data files half the size of the LIF files.
DATA ARRAY ON RAW ARRAY ON FORMAT ARY ON
J
GRAPHICS ON
If you select , the user graphics area is saved. (Refer to the programmer’s guide for information on using display graphics.) The measurement display is not saved with this selection. (Refer to "If You Are Plotting Measurement Results to a Disk Drive" on page 4-13 to plot measurement displays to disk.)
J
DATA ONLY ON
If data array is saved along with any other selected array, the instrument state is not saved, and therefore, cannot be recalled.
J Choose for all applications except CITIfile, S2P, or CAE applications.
J Choose for CITIfile, S2P, and CAE applications or when you want to import the information into a spread sheet using comma-separated values (CSV) format.
ASCII Data Formats
CITIfile (Common Instrumentation Transfer and Interchange file) is an ASCII data format that is useful when exchanging data between different computers and instruments. CITIfiles are always saved when the ASCII format has been selected as shown:
Select one of the following choices:
DATA ONLY ON
RAW ARRAY ON
FORMAT ARY ON
SAVE USING ASCII
RETURN
SAVE STATE
Saving Measurement Results · CITIfile · Save/Recall

Printing, Plotting, and Saving Measurement Results Saving Measurement Results
If , or , or is selected, a CITIfile is saved for each displayed channel with the suffix letter “D”, or “F”, followed by a number. The number following “D” and “F” files is the channel number. When is selected, an “r1” file is saved for channel 1/channel 3, and an “r5” file is saved for channel 2/channel 4. For more information on the CITIFile data format as well as a list of CITIFile keywords, refer to the “Understanding the CITIFile Data Format” chapter in the reference guide.
DATA ARRAY ON DATA ONLY ON FORMAT ARY ON
This format creates component data files that describe frequency dependent linear network parameters for 2 port components. These files are assigned a file name with the suffix "S" and are only output (that is, they cannot be read in by the analyzers).
Up to two S2P files are saved: S1 for channel 1, and S2 for channel 2. S2P files are not stored for channel 3 or channel 4 because the data would be redundant. Each S2P file contains all four S-parameter data.
An S2P file is only output when the all of following conditions are met:
DATA ARRAY ON DATA ONLY ON DEFINE DISK SAVE
Error-corrected data CITI files are always saved along with S2P files.
The template for component data files is as follows:
! comment line #
! indicates that all following on this line is a comment
# indicates that entries following on this line are parameters that are being specified
frequency units GHz, MHz, kHz, Hz
parameter S for S-parameters
format DB for dB magnitude and angle in degrees
MA for linear magnitude and angle in degrees
RI for real and imaginary pair
Rn the reference impedance in ohms for the analyzer making the measurement (R 50 or R 75)

Printing, Plotting, and Saving Measurement Results
The "format" choice is selected by the current selection under the FORMAT menu. To select the DB format, the FORMAT must be LOG MAG. For MA, the FORMAT must be LIN MAG (unlike CITIfile), and all other FORMAT selections will output RI data.
The S2P data will always represent the format array data, including effects of electrical delay and port extensions. A CITIfile will be saved at the same time. To be consistent with previous versions, the CITIfile data saved will represent the DATA array (corrected data) without effects of electrical delay or port extensions.
Here is an S2P example file for an 11 point measurement of a 20 dB attenuator:
# HZ S DB R 50 ! Network Analyzer HP8753E.0611
50000000 −56.74 15.178 −20.219 −2.0132 −20.15 −1.6658 −36.188 −123.52
250000000 −53.015 1.7331 −20.373 −10.241 −20.377 −10.029 −33.974 −40.215
450000000 −52.094 5.8173 −20.391 −18.555 −20.387 −17.96 −31.287 61.778
650000000 −51.758 8.02 −20.189 −26.18 −20.112 −26.061 −29.427 153.37
850000000 −50.95 11.472 −20.163 −34.743 −20.198 −34.195 −24.719 −137.83
1050000000 −50.235 9.3562 −20.178 −42.682 −20.19 −42.289 −25.102 −81.096
1250000000 −49.883 9.2574 −20.142 −50.854 −20.223 −50.407 −27.582 −25.509
1450000000 −48.477 5.9944 −20.201 −58.917 −20.21 −58.436 −33.828 35.237
1650000000 −48.462 3.5156 −20.161 −67.008 −20.188 −66.587 −44.184 62.912
1850000000 −47.503 1840 −20.15 −74.862 −20.208 −74.616 −36.893 35.384
2050000000 −46.938 −5.6538 −20.167 −83.048 −20.256 −82.874 −30.385 74.001
Saving Measurement Results

Printing, Plotting, and Saving Measurement Results Saving Measurement Results
Textual measurement results can be saved in a comma-separated value (CSV) format and imported into a spreadsheet application. Additional information is also saved as a preamble to the measurement results. The saved information includes:
Save/Recall SAVE FILE FORMATS
The text file may be retrieved from the floppy disk on personal computer and can be imported into an application that accepts text in the comma-separated value format, such as a spreadsheet.

Printing, Plotting, and Saving Measurement Results
How the Analyzer Names These Files Sequentially
When text files are saved, the analyzer generates the file names automatically in the following format:
txtcss.csv
c is the indicator of the channel (1−4) on which the measurement data was taken.
ss is a 2-digit, sequential indicator of the measurement (file index number). The file index number may be numbered from 00−31. As the next measurement is taken, the file index number is incremented.
If all four channels are making measurements and a save is performed, there will be four channel numbers that share the same file index number. For example, the files would be named txt100.csv, txt200.csv, txt300.csv, and txt400.csv. If a measurement does not include all four channels, unused channel-file index numbers will not be used by the next measurement. However, if all of the files that share a file index number are erased, that file index number will be re-used.
csv is the file format, comma-separated value in this case.
Saving Measurement Results

Printing, Plotting, and Saving Measurement Results Saving Measurement Results
Graphical measurement results can be saved in JPEG format and used as an illustration in a text editor or desktop publishing application.1 Up to eight traces may be saved in the JPEG file. This is done by storing a measurement using and turning on for each of the four channels.
Save/Recall SAVE FILE FORMATS
The graphic file may be retrieved from the floppy disk on personal computer and can be imported into an application that accepts graphics in the JPEG format.
You may change the analyzer to the factory default color settings by pressing
Copy DEFINE PRINT DEFAULT PRNT SETUP
to correct this problem.
However, to maintain your current color settings (except white), check the measurement color settings by pressing . Press to check the remaining measurement colors. To modify any of the measurement colors, select the measurement and then choose another color from the list of colors that is displayed.
Copy DEFINE PRINT PRINT COLORS MORE

Printing, Plotting, and Saving Measurement Results
When an instrument state is saved to a floppy disk, some or all of the following files may be produced. This depends upon which arrays are selected under the softkey menu, and whether the selected save format is BINARY or ASCII. The “XX” part of the file name (FileXX) refers to the number of the instrument state. The first instrument state saved to any particular disk will be named “File00”, and each successive state saved to that disk will be numbered “1” higher than the previous state (for example, “File01”and “File02”).
Files with .i and .p File Extensions
The following two files (.i file and .p file) are always produced, except when is selected. These files were separated to allow backward compatibility with older instruments. The binary data contained in these two files is not meant to be read in an external computer:
Files with .10, .11, .12, .1a, .1b, and .1c File Extensions
The following files are only produced if you have an active calibration. FileXX.10 is a binary file which stores the stimulus state of the instrument as it relates to an active calibration (specifically, the Power, Sweep Setup, Start, Stop, Center, and Span settings).
The same type of file is produced if Channel 2 is active, but the file extension is .20 instead of the .10 file extension mentioned in the previous paragraph.
Files FileXX.11 through .12, .1a, .1b, and .1c are binary files which contain the 12 error correction coefficients for Channel 1. If Channel 2 is active, it will have the same array, but file extensions are in the form .21, .22, .2a., .2b, and .2c. If you save in ASCII format, only .10 and .1c are produced, with .1c containing the entire error correction array in a two-column, real/imaginary (CITIfile) format.
Files with File Extensions .r1 through .r8
FileXX.r1 through .r4 are produced when is turned ON. They may be either binary or ASCII, and contain the four raw uncorrected S-parameters for channel 1. Channel 2 has the same array, but file extensions are .r5 through .r8. In ASCII format, the data is displayed as two columns of real/imaginary numbers (CITIfile format); S11 appears first, S21 appears second, S12 appears third, and S22 appears last.
RAW ARRAY on OFF
Saving Measurement Results

Printing, Plotting, and Saving Measurement Results Saving Measurement Results
Files with .d1 and .d2 File Extensions
There are two type of files with .d1 and .d2 file extensions. There is FileXX.d1 (or .d2) and DataXX.d1 (or .d2).
FileXX.d1, produced only when is turned ON, may be either binary or ASCII. This file contains the error-corrected measurement data, but without port extensions or electrical delay. In ASCII format, this is a two-column real/imaginary array (CITIfile format), without any direct frequency information. S11 appears first, S21 second, S12 third, and S22 last. If Channel 2 is active, the same type of file is produced but the file extension is .d2. If dual display is on, both d1 and .d2 files are produced.
DATA ARRAY on OFF
DataXX.d1, created only when is turned ON, is either a binary file or an ASCII file in CITIfile format. Turning the softkey ON suppresses the generation of all the previous file types. The contents of this file are identical to those of the FileXX.d1 file. The same type of file is produced if Channel 2 is active, but the file extension is .d2. If dual display is on, both .d1 and .d2 files are produced.
DATA ONLY on OFF
DATA ONLY on OFF
NOTE The DataXX files are much smaller than an entire instrument state, and are the best way to get just the data you want without saving the entire instrument state. Selecting more than one disk-save option does not confuse the analyzer, and simply produces all files associated with the selected options. The only exception to this is that selecting suppresses all other selected options.
DATA ONLY on OFF
Files with .s1 and .s2 File Extensions
There are two type of files with .s1 and .s2 file extensions. There is FileXX.s1 (or .s2) and DataXX.s1 (or .s2).
DATA ONLY on OFF
FileXX.s1 is an ASCII file in Touchstone S2P format. Basically, this is a file in real/imaginary spreadsheet type format with five columns: frequency in the first column, S11 in the second column, S21 in the third, S12 in the fourth, and S22 in the fifth column. If Channel 2 is active, the same type of file is produced but the file extension is .s2. If dual display is on, both .s1 and .s2 are produced. These Touchstone S2P files are only produced when a full 2-port calibration is active and is selected. The effects of port extensions and electrical delay, if they are turned on, are included in the data.
DATA ONLY ON off
DataXX.s1 is also an ASCII file in Touchstone S2P format. As with FileXX.s1, DataXX.s1 is a five-column real/imaginary spreadsheet type format where the columns are used the same as FileXX.s1. If Channel 2 is active, the same type of file is produced but the file extension is .s2. If dual display is on, both .s1 and .s2 are produced. These Touchstone S2P files are only produced when a full 2-port calibration is active and is selected. The effects of port extensions and electrical delay, if they are turned on, are included in the data.
With Turned Off · With Turned On

Printing, Plotting, and Saving Measurement Results
Files with .g0 File Extension
FileXX.g0, produced only when is turned ON, is a binary file containing the active measurement trace and display graticule. The contents of this file are not meant to be read in an external computer, so this file is only of use in the instrument.
The size of the data files is very small, about one tenth the size compared to the ASCII format. Binary is the format to use when you want to store and recall instrument states on the analyzer quickly, but do not need to read the data in an external computer.
Viewing Files Within the Analyzer
All these files are rolled up into a single instrument state, so the analyzer shows only the “FileXX” part of the name, without an extension. The file description will say ISTATE, followed by parentheses with letters in them, such as (CDG). These letters are explained on the bottom of the analyzer screen, and indicate some of what is included in that instrument state. In this example, the state includes the Calibration Array, Data, and Graphics.
The only way to see all the file extensions previously described is to save the instrument state to a disk and view the file structure on an external computer.
Internal data processing is done sequentially beginning with raw data and ending with error correction and all formatting applied. The time domain processing occurs near the end of this processing chain, so data showing the effects of time domain processing is only available in "formatted arrays".
The following discussion explains the data processing flow in the network analyzer. This information is very important if you will be utilizing data from your analyzer for use in computer applications such as spreadsheets, word processing, programs, etc. Refer to Figure 4-13 on page 4-40.
The analyzer receives data from its A, B, and R or Aux-in inputs. Notice the three highlighted blocks. They represent the raw, data, and format arrays. You can save your measurement data in any or all of these format arrays each time the data is saved. Select the arrays of interest based on the factors discussed in this section.
For this discussion, only DISK saves will be described. Data can be saved to internal non-volatile memory or transferred over GPIB as well as to a floppy disk. You will find multiple files saved depending on the arrays chosen under the analyzer's
menu. When using these files, it is important to know which file extension is needed for your particular job.
Saving Measurement Results · GRAPHICS on OFF · Binary Files

Printing, Plotting, and Saving Measurement Results Saving Measurement Results
On the analyzer, press the
Save/Recall DEFINE DISK-SAVE RAW ARRAY ON
Data created the first time in this manner will be saved as filename “FILE00.r1”. The file extension .r1 indicates the data was created while channel 1 was active and stored in the analyzer's raw data array. If you save the data again, but while channel 2 is active, you will get a new file called “FILE01.r2”. RAW data are not commonly used unless sophisticated data processing is to be performed in an external PC. As an example, multi-port calibration is created by exporting raw data to a PC where error-correction for each of the multi-port paths is applied to them.
Save/Recall DEFINE DISK-SAVE DATA ARRAY ON
Press .
Data created the first time in this manner will be saved as filename “DATA00.d1.” The file extension .d1 indicates that the data from the analyzer's channel 1 is error corrected data only if the analyzer's error correction feature is enabled (in other words, you have performed a calibration). Otherwise, the data is the same as data stored in the analyzer's raw data arrays. Data stored in the data arrays does not have any formatting applied to it.
Save/Recall DEFINE DISK-SAVE FORMAT ARY ON
Data created the first time in this manner will be saved as filename “FILE00.f1”. The file extension .f1 indicates the data is formatted per Figure 4-13 using the analyzer's channel 1. Depending on what features you've selected, data in the format arrays includes data in the data arrays plus one or more of the following features:
In each of these examples, most users will select , under the softkey menu. If is turned ON, an additional file will be created with file extension .g0. This is a Hewlett-Packard Graphics Language (HPGL) file.
SAVE USING ASCII DEFINE DISK-SAVE GRAPHICS on OFF

Printing, Plotting, and Saving Measurement Results
Re-Saving an Instrument State
If you re-save a file, the analyzer overwrites the existing file contents.
INTERNAL MEMORY
J
INTERNAL DISK
J (If necessary, refer to the external disk setup procedure in "Saving an Instrument State" on page 4-38.)
Save/Recall SELECT DISK

Printing, Plotting, and Saving Measurement Results Deleting a File
INTERNAL MEMORY
J
INTERNAL DISK
J (If necessary, refer to the external disk setup procedure in "Saving an Instrument State" on page 4-38.)
To Delete an Instrument State File
J Press the or keys or the front panel knob to highlight the name of the file that you want to delete.
J Press to delete all of the files that make up the selected instrument state.
To Delete all Files
J Press to delete all of the files that are on the selected storage device.
Save/Recall SELECT DISK

Printing, Plotting, and Saving Measurement Results
INTERNAL MEMORY
J
INTERNAL DISK
J (If necessary, refer to the external disk setup procedure in "Saving an Instrument State" on page 4-38.)
DONE
Renaming a File · Save/Recall AUTO-FEED OFF

Printing, Plotting, and Saving Measurement Results Recalling a File
INTERNAL MEMORY
J
INTERNAL DISK
J (If necessary, refer to the external disk setup procedure in "Saving an Instrument State" on page 4-38.)
Formatting a Disk
Save/Recall FILE UTILITIES FORMAT DISK
FORMAT:LIF
J
FORMAT:DOS
Save/Recall SELECT DISK

Printing, Plotting, and Saving Measurement Results Solving Problems with Saving or Recalling Files
If you encounter a problem when you are storing files to disk, or the analyzer internal memory, check the following list for possible causes:
Local SYSTEM CONTROLLER

Printing, Plotting, and Saving Measurement Results Solving Problems with Saving or Recalling Files


Optimizing Measurement Results Using This Chapter
This chapter describes techniques and analyzer functions that help you achieve the best measurement results. The following topics are included in this chapter:
J Improper calibration techniques
J Sweeping too fast for electrically long devices
J Reference plane and port extensions

Optimizing Measurement Results Taking Care of Microwave Connectors
Proper connector care and connection techniques are critical for accurate, repeatable measurements.
Refer to the calibration kit documentation for connector care information. Prior to making connections to the network analyzer, carefully review the information about inspecting, cleaning and gaging connectors.
Having good connector care and connection techniques extends the life of these devices. In addition, you obtain the most accurate measurements.
This type of information is typically located in chapter 3 of the calibration kit manuals.
For additional connector care instruction, contact your local Agilent Technologies Sales and Service Office about course numbers HP/Agilent 85050A+24A and 85050A+24D.
See the following table for quick reference tips about connector care.
Table 5-1 Connector Care Quick Reference
Do Keep connectors clean Do Not Touch mating-plane surfaces
Extend sleeve or connector nut Set connectors contact-end down
Use plastic end-caps during storage
Do Inspect all connectors carefully Do Not Use a damaged connector - ever
Look for metal particles, scratches, and dents
Do Try compressed air first Do Not Use any abrasives
Use isopropyl alcohol Get liquid into plastic support beads
Do Clean and zero the gage before use Do Not Use an out-of-spec connector
Use the correct gage type
Use correct end of calibration block
Gage all connectors before first use
Do Align connectors carefully Do Not Apply bending force to connection
Make preliminary connection lightly Over tighten preliminary connection
Turn only the connector nut Twist or screw any connection
Use a torque wrench for final connect Tighten past torque wrench "break" point

Optimizing Measurement Results Increasing Measurement Accuracy
The following all contribute to loss of accuracy in a measurement.
Cables that connect the device under test (DUT) to the analyzer are often the most significant contribution to random errors of your measurement. You should frequently perform the following steps as a precaution against errors caused by cable interconnections:
Calibrations techniques performed improperly contribute to random errors to your measurement. You should frequently perform the following steps as a precaution against errors caused by calibration techniques:
It is possible to sweep too fast for electrically long devices. This will result in measurement error. Refer to "Making Accurate Measurements of Electrically Long Devices" on page 5-8.
Connector repeatability is a source of random measurement error. Measurement error-corrections do not compensate for these errors. For all connectors, you should frequently perform the following steps as a precaution against errors caused by connector repeatability:

Optimizing Measurement Results Increasing Measurement Accuracy
Electrical characteristics will change with temperature due to the thermal expansion characteristics of devices within the analyzer, calibration devices, test devices, cables, and adapters. Therefore, the operating temperature is a critical factor in their performance. During a measurement calibration, the temperature of the calibration devices must be stable and within 25 ±5 °C.
Minute changes in frequency accuracy and stability can occur as a result of temperature and aging (on the order of parts per million). If you require greater frequency accuracy, override the internal crystal with a high-stability external source, frequency standard, or (if your analyzer is equipped with Option 1D5) use the internal frequency standard.
You should periodically check the accuracy of the analyzer measurements, by performing a measurement verification at least once per year. The service guide includes the measurement verification procedure.
Use the port extension feature to compensate for the phase shift of an extended measurement reference plane, due to such additions as cables, adapters, and fixtures, after completing an error-correction procedure (or when there is no active correction).
Using port extensions is similar to using electrical delay. However, using port extensions is the preferred method of compensating for test fixture phase shift. Table 5-2 explains the difference between port extensions and electrical delay.

Optimizing Measurement Results Increasing Measurement Accuracy
Table 5-2 Differences between PORT EXTENSIONS and ELECTRICAL DELAY
Main Effect The end of a cable becomes the test port plane for all S-parameter measurements.
Compensates for the electrical length of a cable. Set the cable’s electrical length × 1 for transmission. Set the cable’s electrical length × 2 for reflection.
Measurements Affected All S-parameters. Only the currently selected S-parameter.
Electrical Compensation Intelligently compensates for 1 times or 2 times the cable’s electrical delay, depending on which S-parameter is computed.
Only compensates for electrical length.
You can activate a port extension by pressing . Then enter the delay to the reference plane.
Cal MORE PORT EXTENSIONS EXTENSIONS ON

Optimizing Measurement Results Maintaining Test Port Output Power During Sweep Retrace
During standard operation, the analyzer provides output power during its forward frequency sweep, but may not provide output power during its sweep retrace. If the device under test (such as an amplifier with AGC circuitry) requires constant power, then you can set the analyzer to maintain test port output power during its sweep retrace.
To activate this feature, press . The analyzer now maintains test port output power during sweep retrace.
System CONFIGURE RETRCE PWR ON

Optimizing Measurement Results Making Accurate Measurements of Electrically Long Devices
A device with a long electrical delay, such as a long length of cable, a SAW filter, or normal devices measured over wide sweeps with very fast rates presents some unusual measurement problems to a network analyzer operating in swept frequency mode. Often the measured response is dependent on the analyzer’s sweep time, and incorrect data may be obtained. At faster sweep rates, the magnitude of the response may seem to drop and look distorted, while at slower sweep rates it looks correct. The results may indicate that a cable has more loss than it truly does, or that a filter has some unusual ripple in the passband which is not really there.
This section describes the cause of this behavior, and how to accurately measure these electrically long devices.
When using a vector network analyzer to measure a device that has a long electrical delay (∆T), the device’s time delay causes a frequency shift between its input and output signals. The frequency shift, ∆F, equals the product of the sweep rate and the time delay:
∆F = dF/dt × ∆T
Since frequency is changing with time as the analyzer sweeps, the time delay of the DUT causes a frequency offset between its input and output. In the analyzer receiver, the test and reference input signals will differ in frequency by ∆F. Because the test signal frequency is slightly different than the receiver frequency, the analyzer will err in measuring its magnitude or phase. The faster the analyzer’s sweep rate, the larger ∆F becomes, and the larger the error in the test channel.
The analyzer does not sweep at a constant rate. The frequency range is covered in several bands, and the sweep rate may be different in each band. So if an operator sets up a broadband sweep with the minimum sweep time, the error in measuring a long device will be different in each band, and the data will be discontinuous at each band edge. This can produce confusing results which make it difficult to determine the true response of the device.
To reduce the error in these measurements, the frequency shift, ∆F, must be reduced. ∆F can be reduced by using the following methods:

Optimizing Measurement Results Making Accurate Measurements of Electrically Long Devices
Decreasing the Sweep Rate
The sweep rate can be decreased by increasing the analyzer’s sweep time. To increase the analyzer’s sweep time, press and use the front panel knob, the and keys, or the front panel keypad enter in the appropriate sweep time.
Sweep Setup SWEEP TIME [MANUAL]
Alternatively, the number of points may be increased for the same frequency range, thereby reducing the sweep rate (in GHz/second).
Selection of the appropriate sweep time depends on the device being measured; the longer the electrical delay of the device under test, the slower the sweep rate must be. A good way to tell when the sweep rate is slow enough is to put the vector network analyzer into a stepped mode of sweeping ( ), and compare the data. In this mode, the vector network analyzer does not sweep the frequency, but steps to each frequency point, stops, makes a measurement, then goes on to the next point. Because errors do not occur in the stepped frequency mode, it can be used to check the data. The disadvantage of the stepped frequency mode is that it is slower than sweeping.
Sweep Setup SWEEP TYPE MENU STEP SWP ON off
Decreasing the Time Delay
The other way to reduce ∆F is by decreasing the time delay, ∆T. Since ∆T is a property of the device that is being measured, it cannot literally be decreased. However, what can be decreased is the difference in delay times between the paths to the R channel and the B channel. These times can be equalized by adding a length of cable to the R channel which has approximately the same delay as the device under test.
This length of cable can be inserted between the R CHANNEL IN and OUT connectors on the front panel of the analyzer. The delay of this cable must be less than 5µs.
![Page 286 of the Agilent Technologies Network Analyzer (20ES) user manual — DONE LIST TYPE: [SWEPT]](/cdn/reader/15/150a289e983abebdc4bd6f0c3204b03f577bfd829d08c0cf5bb1353a14e103d7/images/page-286.webp?w=400)
Optimizing Measurement Results Increasing Sweep Speed
You can increase the analyzer sweep speed by avoiding the use of some features that require computational time for implementation and updating, such as bandwidth marker tracking.
You can also increase the sweep speed by making adjustments to the measurement settings. The following suggestions for increasing sweep speed are general rules that you should experiment with:
When using a list frequency sweep, choosing swept list mode can increase throughput by up to 6 times over stepped list mode. This mode takes data while sweeping through each list segment. In addition, this mode expands the list table to include test port power and IF bandwidth. Selectable IF bandwidths can increase the throughput of the measurement by allowing the user to specify narrow bandwidths only where needed.
Sweep Setup SWEEP TYPE MENU EDIT LIST ADD

IF delay occurs during swept measurements when the signal from the analyzer source is delayed in reaching the analyzer receiver because of an electrically long device. The receiver has a narrow IF band pass filter that tracks the receiver frequency because the receiver is sweeping. The delayed signal will be attenuated because the center of the internal IF filter has moved.
For most measurements, swept list mode will be the optimum choice. If there is any doubt about the effect of IF delay, perform the following test:
Sweep Setup SWEEP TYPE MENU EDIT LIST LIST TYPE: [STEPPED] DONE
The hardware of the network analyzer sweeps the frequency range in separate bands, where switching from band to band takes time. Modify the frequency span to eliminate as many band switches as possible while maintaining measurement integrity. Refer to the following table to identify the analyzer's band switch points:
Table 5-3 Band Switch Points
Band Frequency Span Band Frequency Span
1 50 MHz to 110 MHz 8 2.55 GHz to 4.71 GHz
2 110 MHz to 230 MHz 9 4.71 GHz to 8.256 GHz
3 230 MHz to 470 MHz 10 8.256 GHz to 13.562 GHz
4 470 MHz to 698 MHz 11 (8720ET/ES and 8722ET/ES)
13.562 GHz to 20.05 GHz
5 698 MHz to 1.17 GHz 12 (8722ET/ES) 20.05 GHz to 25.0 GHz
6 1.17 GHz to 1.878 GHz 13 (8722ET/ES) 25.0 GHz to 40.0 GHz
7 1.878 GHz to 2.55 GHz

Optimizing Measurement Results Increasing Sweep Speed
Auto sweep time mode is the default mode (the preset mode). This mode maintains the fastest sweep speed possible for the current measurement settings.
Sweep Setup SWEEP TIME 0
The specifications and characteristics chapter of the reference guide shows the relative increase in sweep time as you decrease system bandwidth.
By reducing the averaging factor (number of sweeps) or switching averaging off, you can increase the analyzer's measurement speed. The time needed to compute averages can also slow the sweep time slightly, in narrow spans.
Menu NUMBER OF POINTS
Refer to the “Specifications and Characteristics” chapter of the reference guide for examples of how sweep time changes with the number of points.
Different sweep speeds are associated with the following three types of non-power sweeps. Choose the sweep type that is most appropriate for your application.
Sweep Setup SWEEP TYPE MENU
Avg IF BW · Avg AVG FACTOR · Avg AVERAGING OFF

Viewing a single channel will increase the measurement speed if the analyzer’s channels are in alternate, or uncoupled mode.
Display DUAL | QUAD SETUP DUAL CHAN on OFF AUX CHAN on OFF
Chan 1 Chan 2
If you must view both measurement channels simultaneously (with dual channel), use the chop sweep mode, explained next.
Chan 3 Chan 4
Display DUAL | QUAD SETUP AUX CHAN on OFF
To return to a single-trace display, press .
You can use the chop sweep mode to make two measurements at the same time. For example, the analyzer can measure S11 and S21 simultaneously. You can activate the chop mode by pressing or by pressing .
Preset Cal MORE CHOP A and B
While Chop mode is the fastest way to measure devices, some components, such as filters with very high attenuation, may require measurement in Alternate mode. See "Increasing Dynamic Range" on page 5-15.
Off-loading the error correction process to an external PC increases throughput on the network analyzer. This can be accomplished with remote-only commands. Refer to the programmer’s guide for information on how to use external calibration.
With a 2-port calibration on, faster trace updates are possible by not measuring the reverse path for every forward sweep. This is controlled by the test set switch command. This is convenient for tuning applications because it gives a faster trace update.
When making measurements using full two-port error-correction, the following types of test set switching can be defined by the user:
Optimizing Measurement Results · Increasing Sweep Speed · Meas

Optimizing Measurement Results Increasing Sweep Speed
For most devices, the fast 2-Port Calibration method is nearly as accurate as the full 2-Port Calibration method.
Cal MORE TESTSET SW CONTINUOUS
The analyzer will then display the softkey as .
The analyzer will then display .
The analyzer will then display the softkey as .
TESTSET SW 8 Sweeps

Dynamic range is the difference between the analyzer’s maximum allowable input level and minimum measurable power. For a measurement to be valid, input signals must be within these boundaries. The dynamic range is affected by these factors:
You can increase the analyzer's source output power so that the test device output power is at the top of the measurement range of the analyzer test port.
Press and enter the new source power level, followed by .
Refer to "Reducing Noise" on page 5-16.
Refer to "Reducing Receiver Crosstalk" on page 5-18.
Optimizing Measurement Results · Increasing Dynamic Range · Power x1

Optimizing Measurement Results Reducing Noise
You can use two analyzer functions to help reduce the effect of noise on the data trace:
The noise is reduced with each new sweep as the effective averaging factor increments.
Refer to "Averaging" on page 7-34 for more information.
By reducing the system bandwidth, you reduce the noise that is measured during the sweep. While averaging requires multiple sweeps to reduce noise, narrowing the system bandwidth reduces the noise on each sweep, however, the sweep will be slower.
Narrower system bandwidths cause longer sweep times. When in auto sweep time mode, the analyzer uses the fastest sweep time possible for any selected system bandwidth. Auto sweep time mode is the default (preset) analyzer setting.
If your instrument is equipped with Option 085 (High Power System), another way of increasing dynamic range is to increase the input power to the test device using a booster amplifier.
Avg AVERAGING FACTOR · Avg IF BW

Direct sampler access to both the A and B samplers can decrease the noise floor of the analyzer.
In the standard configuration (Option 012 with all jumpers in place), the signal entering one of the test ports passes through a coupler with 14 dB coupling before it reaches the sampler. Thus, the input signal is reduced by approximately 14 dB before it reaches the sampler. By using direct sampler access and inputting the signal from the DUT directly in the A or B sampler input, you can improve the noise floor of the analyzer by 14 dB.
Figure 5-1 Increased Dynamic Range Configuration

Optimizing Measurement Results · Reducing Noise

Optimizing Measurement Results Reducing Receiver Crosstalk
To reduce receiver crosstalk you can do the following:
Alternate sweep is intended for measuring wide dynamic range devices, such as high pass and bandpass filters. This sweep mode removes a type of leakage term through the device under test, from one channel to another.
Cal MORE ALTERNATE A and B
To set the alternate sweep, press .
Refer to "Frequency Response and Isolation Error Corrections" on page 6-17.
To reduce time during recall and frequency changes, the raw offset function can be turned off. To turn this function off, press .
System CONFIGURE MENU RAW OFFSET OFF
The raw offset function is normally on and controls the sampler and attenuator offsets. The creation of this table takes some time during a recall of an instrument state.
Raw offsets may be turned on or off individually for each channel. They follow the channel coupling. For dual channel operation, raw offsets should be turned off for each channel if the channels are uncoupled.
Refer to “Specifications and Characteristics” chapter in the reference guide for examples of recall state times with the following functions on or off: raw offsets and blank display. Using blank display may speed up recall times.


Calibrating for Increased Measurement Accuracy How to Use This Chapter
This chapter is divided into the following subjects:
— frequency response error correction
— frequency response and isolation error correction
— enhanced frequency response error correction (with enhanced reflection error
— one-port reflection error correction
— full two-port error correction (ES Analyzers Only)
— adapter removal calibration (ES Analyzers Only)
— modify the cal kit thru definition

Calibrating for Increased Measurement Accuracy
The accuracy of network analysis is greatly influenced by factors external to the network analyzer. Components of the measurement setup, such as interconnecting cables and adapters, introduce variations in magnitude and phase that can mask the actual response of the device under test.
Error correction is an accuracy enhancement procedure that removes systematic errors (repeatable measurement variations) in the test setup. The analyzer measures known standard devices, and uses the results of these measurements to characterize the system.
Measurement accuracy and system characteristics can be affected by the following factors:
If your test setup meets any of the these conditions, the following system characteristics may be affected:
Introduction

Calibrating for Increased Measurement Accuracy Calibration Considerations
Calibration procedures are parameter-specific, rather than channel-specific. When a parameter is selected, the instrument checks the available calibration data, and uses the data found for that parameter. For example, if a transmission response calibration is performed for B/R, and an S11 1-port calibration for A/R, the analyzer retains both calibration sets and corrects whichever parameter is displayed. Once a calibration has been performed for a specific parameter or input, measurements of that parameter remain calibrated in either channel, as long as stimulus values are coupled. In the response and response and isolation calibrations, the parameter must be selected before calibration. Other correction procedures select parameters automatically. Changing channels during a calibration procedure invalidates the part of the procedure already performed.
In calibration procedures that require measurement of several different devices, for example a short, an open, and a load, the order in which the devices are measured is not critical. Any standard can be re-measured, until the key is pressed. The change in trace during measurement of a standard is normal.
“Response” and “response and isolation” calibrations require measurement of only one standard device. If more than one device is measured, only the data for the last device is retained.
When you are performing error correction for a system that has type-N port connectors, the softkey menus label the sex of the test port connector — not the calibration standard connector. For example, the label, SHORT (F), refers to the short that will be connected to the female test port. Since many devices have type-N (f) connectors, the calibration standard to select is the type-N (m). Be sure to use the port extension pin on the type-N (m) Open calibration standard.
Isolation calibration can be omitted for most measurements, except where high dynamic range is a consideration. Use the following guidelines. When the measurement requires a dynamic range of:

Calibrating for Increased Measurement Accuracy
You should save the calibration data, either in the internal non-volatile memory or on a disk. If you do not save it, it will be lost if you select another calibration procedure for the same channel, or if you change stimulus values. Instrument preset, power on, and instrument state recall will also clear the calibration data.
If you interrupt a calibration to go to another menu, such as averaging, you can continue the calibration by pressing the softkey in the correction menu.
The Calibration Standards
During measurement calibration, the analyzer measures actual, well-defined standards and mathematically compares the results with ideal "models" of those standards. The differences are separated into error terms which are later removed during error correction. Most of the differences are due to systematic errors—repeatable errors introduced by the analyzer, test set, and cables—which are correctable.
The standard devices required for system calibration are available in compatible calibration kits with different connector types. Each kit contains at least one short circuit, one open circuit, and an impedance-matched load. In kits that require adapters for interface to the test set ports, the adapters are phase-matched for calibration prior to measurement of non-insertable and non-reversible devices. Other standard devices can be used by specifying their characteristics in a user-defined kit, as described in "Modifying Calibration Kits" on page 7-57.
The accuracy improvement of the correction is limited by the quality of the standard devices, and by the connection techniques used. For maximum accuracy, ensure that the connectors are clean and use a torque wrench for final connections.
Electronic calibration (ECal) modules serve as the calibration standards for electronic calibration procedures. ECal modules are electronic networks that simulate impedance states that are similar to mechanical standards. The factory error correction for optimum performance is embedded in the module’s memory.
Calibration Considerations

Calibrating for Increased Measurement Accuracy Calibration Considerations
In order for the response of a reference standard to show as a dot on the smith chart display format, it must have no phase shift with respect to frequency. Standards that exhibit such "perfect" response are the following:
There are two reasons why other types of reference standards show phase shift after calibration:
The most important point to remember is that these properties will not affect your measurements. The analyzer compensates for them during measurement. As a result, if these standards are measured after a calibration, they will not appear to be "perfect" shorts or opens. This is an indication that your analyzer is working properly and that it has successfully performed a calibration. Figure 6-1 shows sample displays of various calibration standards after calibration.
Some standards have reference planes that are electrically offset from the mating plane of the test port. These devices will show a phase shift with respect to frequency. Table 6-1 shows which reference devices exhibit an electrical offset phase shift. The amount of phase shift can be calculated with the formula:
Φ = (360° × f × l)/c where:
l = electrical length of the offset
c = speed of light (3 x 108 meters/second)
All open circuit terminations exhibit a phase shift over frequency due to fringe capacitance. Offset open circuits have increased phase shift because the offset acts as a small length of transmission line. Refer to Table 6-1.
Electrical Offset · f = frequency · Fringe Capacitance

Calibrating for Increased Measurement Accuracy
Table 6-1 Calibration Standard Types and Expected Phase Shift
Test Port Connector Type Standard Type Expected Phase Shift
7-mm Short 180°
3.5-mm male Offset Short
180° 360° f l × × ( ) c +
3.5-mm female
2.4-mm male
2.4-mm female
75Ω Type-N female
0° Φ°capitance +
3.5-mm male Offset Open
0° Φ°capitance 360° f l × × ( ) c + +
3.5-mm female
2.4-mm male
2.4-mm female
75Ω Type-N female

Calibrating for Increased Measurement Accuracy Calibration Considerations
Figure 6-1 Typical Responses of Calibration Standards after Calibration
3 000 000 000 MHz .
3 000 000 000 MHz .
.300 000 MHZ
STOP 3
000 000 000 MHZ .
.300 000 MHZ
STOP 3
000 000 000 MHZ .
7mm or Type-N Male Short (No Offset)
Type-N Female, 3.5mm Male or Female Offset Short
3 000 000 000 MHz .
3 000 000 000 MHz .
.300 000 MHZ
STOP 3
000 000 000 MHZ .
.300 000 MHZ
STOP 3
000 000 000 MHZ .
7mm or Type-N Male Open (No Offset) with Fringing Capacitance
Type-N Female 3.5 mm Male or Female Offset Open
You may want to use interpolated error correction when you choose a subset of a frequency range that you already corrected, when you change the number of points, or when you change to CW. This feature also allows you to change the parameters in a 2-port correction, such as IF bandwidth, power, or sweep time. The analyzer calculates the systematic errors from the errors of the original correction.
Cal INTERPOL ON off CORRECTION on OFF
To activate interpolated measurement correction, press so that ON is selected, and so that ON is selected. When interpolation is in use, the notation C∆ will appear on the analyzer display.

Calibrating for Increased Measurement Accuracy
System CONFIGURE MENU USER SETTINGS PRESET SETTINGS CAL INTERP ON off
System performance is unspecified when using interpolated error correction. The quality of the interpolated error correction is dependent on the amount of phase shift and the amplitude change between measurement points. If phase shift is no greater than 180° per approximately five measurement points, interpolated error correction offers a great improvement over uncorrected measurements. The accuracy of interpolated error correction improves as the phase shift and amplitude change between adjacent points decrease. When you use the analyzer in linear frequency sweep, perform the original calibration with at least 30 points per 1 GHz of frequency span for greatest accuracy with interpolated error correction.
Interpolated error correction is available in three sweep modes: linear frequency, power sweep, and CW time.
Error correction is only valid for a specific stimulus state, which you must select before you start a correction. If you change any of the following parameters, you will invalidate the correction and the analyzer will switch the correction off (unless the interpolated error correction feature is activated):
The error correction quality may be degraded (Cor changes to C∆), if you change the following stimulus state parameters:
If correction is turned off or shows C∆, the original stimulus state can be recalled by, first, turning interpolation off ( ), and toggling correction off and then on ( ).
INTERPOL ON off CORRECTION ON off
Calibration Considerations

Calibrating for Increased Measurement Accuracy Procedures for Error Correcting Your Measurements
This section has example procedures or information on the following topics:
Several types of error correction are available that remove from one to twelve systematic errors. The full 2-port correction effectively removes all twelve correctable systematic errors. Some measurements do not require correction for all twelve errors. Table 6-2 explains each correction and its uses.
Table 6-2 Purpose and Use of Different Error Correction Procedures
Correction Procedure Corresponding Measurement
Errors Corrected Standard Devices
Response Transmission or reflection measurement when the highest accuracy is not required.
Frequency response. Thru for transmission, open or short for reflection.
Response & Isolation Transmission of high insertion loss devices or reflection of high return loss devices. Not as accurate as 1-port or 2-port correction.
Frequency response plus isolation in transmission or directivity in reflection.
Same as response plus isolation standard. (load)

Calibrating for Increased Measurement Accuracy Procedures for Error Correcting Your Measurements
Table 6-2 Purpose and Use of Different Error Correction Procedures
Correction Procedure Corresponding Measurement
Errors Corrected Standard Devices
Enhanced Response and Enhanced Reflection
Transmission or reflection measurement when improved accuracy is desired. Not as accurate as 2-port calibration.
Directivity, source match, and frequency response for reflection. Frequency response, source match, and isolation for transmission. Enhanced reflection corrects for load match on bilateral devices.
Short, open, load, and thru or ECal module.
S11 1-port Reflection of any one-port device or well terminated two-port device.
Directivity, source match, frequency response.
Short, open and load or ECal module.
S22 1-port (ES Analyzers Only)
Reflection of any one-port device or well terminated two-port device.
Directivity, source match, frequency response.
Short, open and load or ECal module.
Full 2-port (ES Analyzers Only)
Transmission or reflection of highest accuracy for two-port devices.
Directivity, source match, load match, isolation, frequency response, forward and reverse.
Short, open, load and thru (2 loads for isolation) or ECal module
TRL*/LRM* (ES Analyzers Only)
Transmission or reflection in noncoaxial environment such as in a fixture or on-wafer.
Directivity, isolation, frequency response. (forward and reverse)
Thru, reflect, line, or line, reflect, match, or thru, reflect, match.
TRL/LRM (ES analyzers with Option 400 only)
Transmission or reflection in noncoaxial environments. More accurate than TRL*/LRM*.
Directivity, isolation, frequency response. (forward and reverse)
Thru, reflect, line, or line, reflect, match, or thru, reflect, match.

Calibrating for Increased Measurement Accuracy Frequency Response Error Corrections
You can remove the frequency response of the test setup for the following measurements:
J If you want to make a reflection measurement on PORT 1 (in the forward direction,
S11), leave the instrument default setting.
J For ES analyzers, if you want to make a reflection measurement on PORT 2 (in the
reverse direction, S22), press:
Cal CAL KIT [ ]
Cal CAL KIT SELECT CAL KIT RETURN
(select your type of kit)
If your type of calibration kit is not listed in the displayed menu, refer to "Modifying Calibration Kits" on page 7-57.
Cal CALIBRATE MENU RESPONSE
Connect the short or open calibration standard to the port you selected for the test port (PORT 1 for S11 or PORT 2 for S22 on ES analyzers).
Preset · Cal CALIBRATE MENU

Calibrating for Increased Measurement Accuracy
Frequency Response Error Corrections
Figure 6-2 Standard Connections for a Response Error Correction for

If the calibration kit you selected has a choice between male and female calibration standards, remember to select the sex that applies to the test port and not the standard.
The analyzer displays WAIT - MEASURING CAL STANDARD during the standard measurement. The analyzer underlines the softkey that you selected after it finishes the measurement, and computes the error coefficients.
This completes the response correction for reflection measurements. You can connect and measure your device under test.
Reflection Measurement

Calibrating for Increased Measurement Accuracy Frequency Response Error Corrections
J If you want to make a transmission measurement in the forward direction (S21),
press:
or on ET models:
J For ES analyzers, if you want to make a transmission measurement in the reverse
direction (S12), press:
Cal CALIBRATE MENU RESPONSE
Figure 6-3 Standard Connections for Response Error Correction for

The analyzer displays WAIT - MEASURING CAL STANDARD during the standard measurement. The analyzer underlines the softkey after it measures the calibration standard, and computes the error coefficients.
Preset · Transmission Measurements

Calibrating for Increased Measurement Accuracy
Frequency Response Error Corrections
Receiver calibration provides a frequency response error correction for a non-ratioed measurement that also indicates absolute power in dBm. This calibration is most useful when performed with a power meter calibration. This calibration is only allowed for non-ratioed measurements A, B, and R.
This calibration normalizes the trace to the current reference value. Typically, this reference value is entered to be the same as the current source power.
Set the analyzer test port power to the desired level (−10 dBm in this example) by pressing:
. This calibrates the receiver to the approximate accuracy of the source output power, which is subject to the source power flatness specification.
Power −10 x1
or

Calibrating for Increased Measurement Accuracy Frequency Response Error Corrections
Figure 6-4 Standard Connections for a Receiver Calibration

Meas INPUT PORTS B
For ES analyzers, press . This sets the source at PORT 1.
Cal CALIBRATE MENU RECEIVER CAL −10 x1 TAKE RCVR CAL SWEEP
Calibrations at powers other than 0 dBm are possible. Receiver calibration normalizes the trace to the value set for the reference level. For example, to do a receiver calibration at −10 dBm, set the source to −10 dBm, set the reference level to −10 dBm, then perform the receiver calibration.

Calibrating for Increased Measurement Accuracy Frequency Response and Isolation Error Corrections
You can make a response and isolation correction for the following measurements:
This procedure is intended for measurements that have a measurement range of greater than 90 dB.
J If you want to make a transmission measurement in the forward direction (S21),
press:
or on ET models:
J For ES analyzers, if you want to make a transmission measurement in the reverse
direction (S12), press:
CAL KIT SELECT CAL KIT RETURN
(select your type of kit)
If your type of calibration kit is not listed in the displayed menu, refer to "Modifying Calibration Kits" on page 7-57.
Preset · Cal

Calibrating for Increased Measurement Accuracy Frequency Response and Isolation Error Corrections
The analyzer displays WAIT - MEASURING CAL STANDARD during the standard measurement. The analyzer underlines the softkey after it measures the calibration standard, and computes the error coefficients.
Figure 6-5 Standard Connections for a Response and Isolation Error
Correction for Transmission Measurements

10.To help remove crosstalk noise, set the analyzer as follows:
Avg AVERAGING ON AVERAGING FACTOR
a. Press and enter at least four times more averages than desired during the device measurement.
b. Press to eliminate one crosstalk path.
Cal MORE ALTERNATE A and B
11.To measure the calibration standard, press:
Cal RESUME CAL SEQUENCE ISOL’N STD

Calibrating for Increased Measurement Accuracy Frequency Response and Isolation Error Corrections
12.Return the averaging to the original state of the measurement. For example, reduce the averaging factor by at least four times or turn averaging off.
13.To compute the isolation error coefficients, press:
Cal RESUME CAL SEQUENCE DONE RESP ISOL’N CAL
The analyzer displays the corrected data trace. The analyzer also shows the notation Cor at the left of the screen, indicating that the correction is switched on for this channel.
14.This completes the response and isolation correction for transmission measurements. You can connect and measure your device under test.
The response and isolation error correction for reflection measurements provides the following benefits:
To perform the response and isolation error correction for reflection measurements:
J If you want to make a reflection measurement on PORT 1 (in the forward direction,
S11), leave the instrument default setting.
J For ES analyzers, if you want to make a reflection measurement on PORT 2 (in the
reverse direction, S22), press:
Preset · Cal

Calibrating for Increased Measurement Accuracy Frequency Response and Isolation Error Corrections
CAL KIT SELECT CAL KIT RETURN
(select your type of kit)
If your type of calibration kit is not listed in the displayed menu, refer to "Modifying Calibration Kits" on page 7-57.
Figure 6-6 Standard Connections for a Response and Isolation Error
Correction for Reflection Measurements

If the calibration kit you selected has a choice between male and female calibration standards, remember to select the sex that applies to the test port and not the standard.
or

Calibrating for Increased Measurement Accuracy Frequency Response and Isolation Error Corrections
The analyzer displays WAIT - MEASURING CAL STANDARD during the standard measurement. The analyzer underlines the softkey that you selected after it finishes the measurement, and computes the error coefficients.
10.To measure the standard for the isolation portion of the correction, press .
a. Press and enter at least four times more averages than desired during the device measurement.
Avg AVERAGING ON AVERAGING FACTOR
11.To compute the response and directivity error coefficients, press:
The analyzer displays the corrected S11 (or S22) data. The analyzer also shows the notation Cor to the left of the screen, indicating that the correction is switched on for this channel.
12.This completes the response and isolation error correction for reflection measurements. You can connect and measure your device under test.
![Page 316 of the Agilent Technologies Network Analyzer (20ES) user manual — CAL KIT [ ]](/cdn/reader/15/150a289e983abebdc4bd6f0c3204b03f577bfd829d08c0cf5bb1353a14e103d7/images/page-316.webp?w=400)
Calibrating for Increased Measurement Accuracy Enhanced Frequency Response Error Correction
The enhanced frequency response error correction removes the following errors in the forward direction in ET models or in both the forward and reverse directions in ES models:
The enhanced reflection error correction may be used to remove load match from the test setup when measuring bilateral devices. Enhanced reflection terms are mathematically derived during all enhanced response calibrations but are not applied unless initiated by the softkey. Enhanced reflection correction is applied after the enhanced frequency response error correction is finished.
ENH. REFL. on OFF
IMPORTANT Use enhanced reflection error correction only on bilateral devices. A bilateral device has similar forward and reverse transmission characteristics. Examples of bilateral devices are passive devices (filters, attenuators, and switches). Most active devices (amplifiers) and some passive devices (isolators and circulators) are not bilateral. If this error correction is used for a non-bilateral device, errors will occur in the resulting measurement.
J If you want to make measurements in the forward direction (S21 S11), press:
or on ET models:
J For ES analyzers, if you want to make measurements in the reverse direction (S12
S22), press:
CAL KIT SELECT CAL KIT RETURN
(select your type of kit)
If your type of calibration kit is not listed in the displayed menu, refer to "Modifying Calibration Kits" on page 7-57.
Preset · Cal

Calibrating for Increased Measurement Accuracy Enhanced Frequency Response Error Correction
J If you want to make measurements in the forward direction, press:
S11/S21 ENH. RESP. TRAN/REFL ENH. RESP.
or on ET models:
J For ES analyzers, if you want to make measurements in the reverse direction, press:
S22/S12 ENH. RESP.
Figure 6-7 Standard Connections for Enhanced Response Calibration

The analyzer displays WAIT - MEASURING CAL STANDARD during the standard measurement. The analyzer underlines the softkey after it measures the standard.
10.To measure the device, when the displayed trace has settled, press:
The analyzer measures the short circuit and underlines the softkey.
11.Disconnect the short, and connect an impedance-matched load to the test port.

Calibrating for Increased Measurement Accuracy Enhanced Frequency Response Error Correction
12.To measure the standard, when the displayed trace has settled, press:
, select the type of load you are using, and then press when the analyzer has finished measuring the load.
13.To compute the reflection correction coefficients, press .
14.To start the transmission portion of the correction, press .
15.Make a "thru" connection between the points where you will connect your device under test as shown in Figure 6-7.
NOTE The thru in most calibration kits is defined with zero length. The correction will not work properly if a non-zero length thru is used, unless the calibration kit is modified to change the defined thru to the length used. This is important for measurements of non-insertable devices (devices having ports that are both male or both female). The modified calibration kit must be saved as the user calibration kit, and the softkey must be selected before the calibration is started.
16.To measure the standard, when the trace has settled, press:
FWD MATCH THRU REV MATCH THRU
STANDARDS DONE
The analyzer underlines the softkey label after it makes each measurement.
17.Press and select from the following two options:
J If you will be measuring devices with a dynamic range less than 90 dB, press:
J If you will be measuring devices with a dynamic range greater than 90 dB, follow
these steps:
a. Connect impedance-matched loads to the test ports. Include the adapters that you
would include for your device measurement.
or · or

Calibrating for Increased Measurement Accuracy Enhanced Frequency Response Error Correction
b. Activate at least four times more averages than desired during the device
measurement.
c. Press .
Cal RESUME CAL SEQUENCE ISOLATION FWD or REV ISOL’N STD DONE
d. Return the averaging to the original state of the measurement, and press .
Cal RESUME CAL SEQUENCE
18.To compute the error coefficients, press .
The analyzer displays the corrected measurement trace. The analyzer also shows the notation Cor at the left of the screen, indicating that error correction is on.
19.If you are measuring a bilateral device and want to remove the load match error, activate the enhanced reflection calibration by pressing:
ENH. REFL. on OFF
until ON is selected
20.This completes the enhanced response correction procedure. You can connect and measure your device under test.

Calibrating for Increased Measurement Accuracy One-Port Reflection Error Correction
You can perform a 1-port correction for an S11 or an S22 measurement for ES analyzers. The only difference between the two procedures is the measurement parameter that you select.
J If you want to make a reflection measurement on PORT 1 (in the forward direction,
S11), leave the instrument default setting.
J For ES analyzers, if you want to make a reflection measurement on PORT 2 (in the
reverse direction, S22), press:
CAL KIT SELECT CAL KIT RETURN
(select your type of kit)
If your type of calibration kit is not listed in the displayed menu, refer to "Modifying Calibration Kits" on page 7-57.
J If you want to make a reflection measurement at PORT 1, press:
J For ES analyzers, if you want to make a reflection measurement at PORT 2, press:
Preset · Cal

Calibrating for Increased Measurement Accuracy
One-Port Reflection Error Correction
Figure 6-8 Standard Connections for a One Port Reflection Error Correction

The analyzer displays WAIT - MEASURING CAL STANDARD during the standard measurement. The analyzer underlines the softkey after it measures the calibration standard.
10.To measure the standard when the displayed trace has settled, press:
The analyzer measures the short circuit and underlines the softkey.
11.Disconnect the short, and connect an impedance-matched load to the test port.
12.When the displayed trace has settled, press:
, select the type of load you are using, and then press when the analyzer has finished measuring the load.
13.To compute the error coefficients, press:

Calibrating for Increased Measurement Accuracy One-Port Reflection Error Correction
The analyzer displays the corrected data trace. The analyzer also shows the notation Cor to the left of the screen, indicating that the correction is switched on for this channel.
14.This completes the one-port correction for reflection measurements. You can connect and measure your device under test.

Calibrating for Increased Measurement Accuracy Full Two-Port Error Correction (ES Analyzers Only)
CAL KIT SELECT CAL KIT RETURN
(select your type of kit)
If your type of calibration kit is not listed in the displayed menu, refer to "Modifying Calibration Kits" on page 7-57.
Cal

Calibrating for Increased Measurement Accuracy Full Two-Port Error Correction (ES Analyzers Only)
Figure 6-9 Standard Connections for Full Two-Port Error Correction
The analyzer displays WAIT - MEASURING CAL STANDARD during the standard measurement. The analyzer underlines the softkey after it measures the standard.
The analyzer measures the short circuit and underlines the softkey.
10.To measure the standard, when the displayed trace has settled, press:
, select the type of load you are using, and then press when the analyzer has finished measuring the load.
11.Repeat the open-short-load measurements described in the previous steps, but connect the devices in turn to PORT 2, and use the , , and softkeys. Include any adapters that you would include in your device measurement.
12.To compute the reflection correction coefficients, press:
13.To start the transmission portion of the correction, press .

Calibrating for Increased Measurement Accuracy Full Two-Port Error Correction (ES Analyzers Only)
14.Make a "thru" connection between the points where you will connect your device under test as shown in Figure 6-9.
NOTE The thru in most calibration kits is defined with zero length. The correction will not work properly if a non-zero length thru is used, unless the calibration kit is modified to change the defined thru to the length used. This is important for measurements of non-insertable devices (devices having ports that are both male or both female). The modified calibration kit must be saved as the user calibration kit, and must be selected before the calibration is started.
15.To measure the standard, when the trace has settled, press:
The analyzer underlines the softkey label after it makes each measurement.
16.Press and select from the following two options:
J If you will be measuring devices with a dynamic range less than 90 dB, press:
J If you will be measuring devices with a dynamic range greater than 90 dB, follow
these steps:
a. Connect impedance-matched loads to PORT 1 and PORT 2. Include the adapters
that you would include for your device measurement.
b. Activate at least four times more averages than desired during the device
measurement.
c. Press .
Cal RESUME CAL SEQUENCE ISOLATION FWD ISOL’N ISOL’N STD REV ISOL’N ISOL’N STD ISOLATION DONE
d. Return the averaging to the original state of the measurement, and press .
Cal RESUME CAL SEQUENCE

Calibrating for Increased Measurement Accuracy Full Two-Port Error Correction (ES Analyzers Only)
17.To compute the error coefficients, press:
The analyzer displays the corrected measurement trace. The analyzer also shows the notation Cor at the left of the screen, indicating that error correction is on.
18.This completes the full two-port correction procedure. You can connect and measure your device under test.

Calibrating for Increased Measurement Accuracy
Power Meter Measurement Calibration
A GPIB-compatible power meter can monitor and correct RF source power to achieve leveled power at the test port. During a power meter calibration, the power meter samples the power at each measurement point across the frequency band of interest. The analyzer then constructs a correction data table to correct the power output of the internal source. The correction table may be saved in an instrument state register with the SAVE key.
The correction table may be updated on each sweep (in a leveling application) or during an initial single sweep. In the sample-and-sweep mode the power meter is not needed for subsequent sweeps. The correction table may be read or modified through GPIB.
Power meter calibration is useful for the following applications:
The power meter can measure and correct power in two ways:
The time required to perform a power meter calibration depends on the source power, number of points tested, and number of readings taken. Refer to the “Specifications and Characteristics” chapter of the reference guide for characteristic power meter calibration sweep speeds and accuracy.
Regardless of the measurement application, the analyzer's source can only supply corrected power within the selected power range. If power outside this range is requested, the annotation will change to PC?.
The power meter calibration data will be lost by committing any of the following actions:
Turning power off. Turning off the instrument erases the power meter calibration table.
Changing sweep type. If the sweep type is changed (linear, log, list, CW, power) while power meter calibration is on, the calibration data will be lost. However, calibration data is retained if you change the sweep type while power meter calibration is off.
Changing frequency. Power meter calibration data will also be lost if the frequency is changed in log or list mode, but it is retained in linear sweep mode.
Pressing. Presetting the instrument will erase power meter calibration data. If the instrument state has been saved in a register using the key, you may recall the instrument state and the data will be restored. Saving the instrument state will not protect the data if the instrument is turned off.
Preset · Save/Recall

Calibrating for Increased Measurement Accuracy Power Meter Measurement Calibration
If the frequency is changed in linear sweep, or the start/stop power is changed in power sweep, then the calibration data is interpolated for the new range.
If calibration power is changed in any of the sweep types, the values in the power setting array are increased or decreased to reflect the new power level. Some accuracy is lost when this occurs.
Entering the power sensor calibration data compensates for the frequency response of the power sensor, thus ensuring the accuracy of power meter calibration.
Cal PWRMTR CAL LOSS/SENSR LISTS CAL FACTOR SENSOR A
The analyzer shows the notation EMPTY, if you have not entered any segment information.
G/n M/µ k/m
You may enter multiple segments in any order because the analyzer automatically sorts them and lists them on the display by frequency value. The analyzer also automatically interpolates the values between correction factor data points.
If you only enter one frequency segment, the analyzer assumes that the single value is valid over the entire frequency range of the correction.
Cal PWRMTR CAL LOSS/SENSR LISTS CAL FACTOR SENSOR A CAL FACTOR SENSOR B
x1 DONE · Editing Frequency Segments · SEGMENT x1


J If you are modifying the frequency, enter the new value, followed by a , , or key.
J If you are modifying the correction factor, enter the new value, followed by the key.
Cal PWRMTR CAL LOSS/SENSR LISTS CAL FACTOR SENSOR A CAL FACTOR SENSOR B
The analyzer deletes the segment and moves the remainder of the segments up one number.
Compensating for Directional Coupler Response
If you use a directional coupler to sample power in your measurement configuration, you should enter the coupled arm power loss value into the power loss table, using the following procedure. You can enter the loss information in a single segment, and the analyzer will assume that the value applies to the entire frequency range of the instrument. Or, you can input actual measured power loss values at several frequencies using up to 12 segments, enhancing power accuracy.
Cal PWRMTR CAL LOSS/SENSR LISTS POWER LOSS
The analyzer shows the notation EMPTY, if you have not entered any segment information.
G/n M/µ k/m
Deleting Frequency Segments · SEGMENT x1 · x1 DONE

Calibrating for Increased Measurement Accuracy Power Meter Measurement Calibration
You may enter multiple segments in any order because the analyzer automatically sorts them and lists them on the display in increasing order of frequency.
If you only enter one frequency segment, the analyzer assumes that the single value is valid over the entire frequency range of the correction.
Cal PWRMTR CAL PWR LOSS ON
You can use the sample-and-sweep mode to correct the analyzer output power and update the power meter correction data table, during the initial measurement sweep. Because the analyzer measures the actual power at each frequency point during the initial sweep, the initial sweep time is significant. However, in this mode of operation the analyzer does not require the power meter for subsequent sweeps. Therefore, this mode sweeps considerably faster than the continuous correction mode.
Figure 6-10 Sample-and-Sweep Mode for Power Meter Calibration


Calibrating for Increased Measurement Accuracy
Power Meter Measurement Calibration
NUMBER OF READINGS n x1
, (where n = the number of desired iterations).
If you increase the number of readings, the power meter correction time will substantially increase.
Cal PWRMTR CAL ONE SWEEP TAKE CAL SWEEP
The analyzer will use the data table for subsequent sweeps to correct the output power level at each measurement point. Also, the status annunciator PC will appear on the analyzer display.
10.Remove the power sensor from the analyzer test port and connect your test device.
Local · Cal PWRMTR CAL · −10 x1

Calibrating for Increased Measurement Accuracy Power Meter Measurement Calibration
You can set the analyzer to update the correction table at each sweep (as in a leveling application), using the continuous sample mode. When the analyzer is in this mode, it continuously checks power at every point in each sweep. You must keep the power meter connected as shown in Figure 6-11. This mode is also known as power meter leveling, and the speed is limited by the power meter.
Figure 6-11 Continuous Correction Mode for Power Meter Calibration

NUMBER OF READINGS n x1
If you increase the number of readings, the power meter correction time will substantially increase.
Cal PWRMTR CAL EACH SWEEP TAKE CAL SWEEP
Cal PWRMTR CAL

Calibrating for Increased Measurement Accuracy
Power Meter Measurement Calibration
To Calibrate the Analyzer Receiver to Measure Absolute Power
You can use the power meter calibration as a reference to calibrate the analyzer receiver to accurately measure absolute power. The following procedure shows you how to calibrate the receiver to any power level.
Cal PWRMTR CAL x1 ONE SWEEP TAKE CAL SWEEP
The status notation PC will appear on the analyzer display. Port 1 is now a calibrated source of power.
Meas INPUT PORTS B TEST PORT 1
This sets the source at PORT 1, and the measurement receiver to PORT 2, or input port B.
Cal CALIBRATE MENU RECEIVER CAL x1 TAKE RCVR CAL SWEEP
The receiver channel now measures power to a characteristic accuracy of 0.35 dB or better. The accuracy depends on the match of the power meter, the source, and the receiver.

Calibrating for Increased Measurement Accuracy Calibrating for Noninsertable Devices
A test device that cannot be connected directly into a transmission test configuration is considered to be noninsertable.
Some examples of noninsertable test devices are:
Therefore, one of the following calibration methods must be performed:
Figure 6-12 Noninsertable Device


Calibrating for Increased Measurement Accuracy
Calibrating for Noninsertable Devices
Adapter removal calibration provides the most complete and accurate procedure for measuring noninsertable devices.
The following adapters are needed:
Figure 6-13 Adapters Needed

Other requirements include:
For each port, a separate 2-port error correction is performed,the first at the connection between A1 and A3, and the second at the connection between A2 and A3. The error coefficients are stored in separate calibration sets. After these calibrations, the two calibration sets are combined and, with knowledge of the electrical length of the adapter A3, a separate third calibration set is created. This cal set contains error coefficients that accurately represent the characteristics of Port 1 and Port 2, as if A1 and A2 were actually connected together to measure forward and reverse match and tracking terms.

Calibrating for Increased Measurement Accuracy Calibrating for Noninsertable Devices
Cal MORE ADAPTER REMOVAL HELP ADAPT REMOVAL
a. Refer to Figure 6-14 while performing the steps in this procedure. Also refer to
page 6-41 for an explanation of A1, A2, and A3.
b. Perform a 1-port calibration at “Reference Port 1”. Refer to Step A of Figure 6-14.
Figure 6-14 Determining the Electrical Delay Setup

c. Connect the A3 adapter to "Reference Port 1" as shown in Step B of Figure 6-14.
Attach the short (from the calibration kit for port 2) to the other end of the adapter.
You must know the delay of the short. The delay of the short can be found in the calibration kit that you are using. Typical delays of shorts are 31.7 ps for the short from the 85052D calibration kit and 31.8 ps for the short from the 85033D calibration kit.
d. Measure the delay of the adapter and short together by pressing .
e. Divide the resulting delay measurement by 2 to determine the delay of the thru and
the short in one direction.
f. Subtract the offset delay of the short (determined in step c) from the delay of the thru and the short in one direction (determined in step e). The result is the electrical delay of the thru. This value is used in the Step 12.
g. Remove the short from the adapter.
Format DELAY

Calibrating for Increased Measurement Accuracy
Calibrating for Noninsertable Devices
Save/Recall SELECT DISK INTERNAL DISK
Figure 6-15 Two-Port Cal Set 1

Figure 6-16 Two-Port Cal Set 2

Save/Recall SAVE STATE

Calibrating for Increased Measurement Accuracy Calibrating for Noninsertable Devices
Cal MORE ADAPTER REMOVAL RECALL CAL SETS
10.Turn the knob to select the file that contains the port 1 calibration data (where adapter A3 was on port 1).
11.Press .
12.Press . Enter the delay value of the adapter from step 2f. Select the appropriate type: or .
ADAPTER COAX ADAPTER WAVEGUIDE
13.Press to complete the technique for calculating the new coeffiecients and overwrite the current active calibration set in use.
This process uses up an internal memory register. The calibration in this register is not the calibration created by adapter removal, rather it is a "scratch" calibration. You may wish to delete the register, or re-save the new calibration in this register (step 14).
14.Press and . You can now remove the adapter from the test setup and insert the DUT.
Save/Recall SELECT DISK INTERNAL MEMORY RETURN SAVE STATE
Figure 6-17 Calibrated Measurement

Save/Recall SAVE STATE

Calibrating for Increased Measurement Accuracy
Calibrating for Noninsertable Devices
Since the effect of the adapter has been removed, it is easy to verify the accuracy of the technique by simply measuring the adapter itself. Because the adapter was used during the creation of the two cal sets, and the technique removes its effects, measurement of the adapter itself should show the S-parameters.
If unexpected phase variations are observed, this indicates that the electrical delay of the adapter was not specified within a quarter wavelength over the frequency range of interest. To correct this, recall both cal sets, since the data was previously stored to disk, change the adapter delay, and press .
Your analyzer’s programmer’s guide contains an example program that demonstrates the adapter removal process over GPIB.
With this method, you use two precision matched adapters which are "equal." To be equal, the adapters must have the same match, Z0, insertion loss, and electrical delay. The adapters in most Agilent calibration kits have matched electrical length, even if the physical lengths appear different.
Figure 6-18 Calibrating for Noninsertable Devices
To use this method, refer to Figure 6-18 and perform the following steps:
The errors remaining after calibration with this method are equal to the differences between the two adapters that are used.

Calibrating for Increased Measurement Accuracy Calibrating for Noninsertable Devices
With this method, it is only necessary to use a thru adapter. The calibration kit thru definition is modified to compensate for the adapter and then saved as a user kit. However, the electrical delay of the adapter must first be found. The adapter match will degrade the effective load match terms on both ports as well as degrade the transmission frequency response (tracking).
Figure 6-19 Determining the Electrical Delay Setup
For example, if A3 has 100 ps of delay, press:
Cal CAL KIT MODIFY DEFINE STANDARD 4 x1
MODIFY STD DEFINITION SPECIFY OFFSET OFFSET DELAY 0.100 G/n
STD DONE (DEFINED) RETURN KIT DONE (MODIFIED) SAVE USER KIT
Format DELAY

Calibrating for Increased Measurement Accuracy
Minimizing Error When Using Adapters
10.Connect the test device as shown in Figure 6-17 and measure the device.
To minimize the error introduced when you add an adapter to a measurement system, the adapter needs to have low SWR or mismatch, low loss, and high repeatability.
Figure 6-20 Adapter Considerations

In a reflection measurement, the directivity of a system is a measure of the error introduced by an imperfect signal separation device. It typically includes any signal that is detected at the coupled port which has not been reflected by the test device. This directivity error will add with the true reflected signal from the device, causing an error in the measured data. Overall directivity is the limit to which a device’s return loss or reflection can be measured. Therefore, it is important to have good directivity to measure low reflection devices.
For example, a coupler has a 7 mm connector and 40 dB directivity, which is equivalent to a reflection coefficient of ρ = 0.01 directivity in dB = −20 log (ρ). Suppose we want to connect to a device with an SMA male connector. We need to adapt from 7 mm to SMA.
If we choose a precision 7 mm to SMA adapter with a SWR of 1.06, which has ρ=0.03, the overall directivity becomes ρ=0.04 or 28 dB. However, if we use two adapters to do the same job, the reflection from each adapter adds up to degrade the directivity to 17 dB. The last example, shown in Figure 6-20, uses three adapters that shows an even worse directivity of 14 dB. It is clear that a low SWR is desirable to avoid degrading the directivity of the system.

Calibrating for Increased Measurement Accuracy Making Non-Coaxial Measurements
Non-coaxial, on-wafer measurements present a unique set of challenges for error correction in the analyzer:
The capability of making non-coaxial measurements is available with TRL (thru-reflect-line) or LRM (line-reflect-match) calibration. For in-depth information on TRL/LRM calibration, refer to "Calibrating for Non-Coaxial Devices (ES Analyzers Only)" on page 6-50.
Due to the simplicity of the calibration standards, TRL or LRM calibrations may be used for non-coaxial applications such as on-wafer measurements. This type of calibration with time-domain gating and a variety of probe styles can provide optimal accuracy in on-wafer measurements.
Fixtures are needed to interface non-coaxial devices to coaxial test instruments. It may also be necessary to transform the characteristic impedance from standard 50 Ω instruments to a non-standard impedance and to apply bias if an active device is being measured.
For accurate measurements, the fixture must introduce minimum change to the test signal, not destroy the test device, and provide a repeatable connection to the device.
For information about test fixtures for your measurement systems, ask for literature number 5962-9723E or contact: Inter-Continental Microwave, 1515 Wyatt Drive, Santa Clara, CA 95054, USA (Web site: http://www.icmicrowave.com).

Calibrating for Increased Measurement Accuracy
If You Want to Design Your Own Fixture
Ideally, a fixture should provide a transparent connection between the test instrument and the test device. This means it should have no loss or electrical length and a flat frequency response, to prevent distortion of the actual signal. A perfect match to both the instrument and the test device eliminates reflected test signals. The signal should be effectively coupled into the test device, rather than leaking around the device and resulting in crosstalk from input to output. Repeatable connections are necessary to ensure consistent data.
Realistically, it is impossible to build an ideal fixture, especially at high frequencies. However, it is possible to optimize the performance of the test fixture relative to the performance of the test device. If the fixture’s effects on the test signal are relatively small compared to the device’s parameters, then the fixture’s effects can be assumed to be negligible.
For example, if the fixture’s loss is much less than the acceptable measurement uncertainty at the test frequency, then it can be ignored.
For additional information about fixtures, refer to Agilent Technologies Application Note 1287-9, “In-Fixture Measurements Using Vector Network Analyzers,” literature number 5968-5329E.
Making Non-Coaxial Measurements

Calibrating for Increased Measurement Accuracy Calibrating for Non-Coaxial Devices (ES Analyzers Only)

The analyzer has the capability of making calibrations using the TRL*/LRM* method. TRL* and LRM* are implementations of the thru-reflect-line and line-reflect-match calibrations, modified for the three-sampler receiver architecture in the analyzer. TRL/LRM calibration is available in analyzers with Option 400, which adds a fourth sampler to the receiver. The following procedures apply to both TRL*/LRM* and TRL/LRM calibrations.
Create a User-Defined TRL Calibration Kit
In order to use the TRL technique, the calibration standards characteristics must be entered into the analyzer’s user defined calibration kit.
The following steps show you how to define a calibration kit to utilize a set of TRL (THRU, REFLECT, LINE) standards. This example TRL kit contains the following:
Modify the Standard Definitions
Preset Cal CAL KIT MODIFY DEFINE STANDARD
1 x1 REFLECT SHORT
SHORT MODIFY STD DEFINITION SPECIFY OFFSET OFFSET DELAY 0
x1 STD OFFSET DONE STD DONE (DEFINED)
DEFINE STANDARD 4 x1 DELAY/THRU MODIFY STD DEFINITION
SPECIFY OFFSET OFFSET DELAY 0 x1 STD OFFSET DONE STD DONE (DEFINED)
DEFINE STANDARD 6 x1 DELAY/THRU MODIFY STD DEFINITION
SPECIFY OFFSET OFFSET DELAY .08 G/n MAXIMUM FREQUENCY
Enter a frequency greater than the maximum frequency range of the analyzer. For example, press . Then, press .
50 G/n STD OFFSET DONE

Calibrating for Increased Measurement Accuracy Calibrating for Non-Coaxial Devices (ES Analyzers Only)
Assign the Standards to the Various TRL Classes
Cal CAL KIT MODIFY SPECIFY CLASS MORE MORE TRL REFLECT
10.Since you previously designated standard #6 for the LINE/MATCH standard, press:
TRL LINE OR MATCH 6 x1
11.Since you previously designated standard #4 for the THRU/LINE standard, press:
TRL THRU 4 x1
12.To complete the specification of class assignments, press:
13.Press .
14.Change the label of the "TRL REFLECT" class to "TRLSHORT."
15.Change the label of the "TRL LINE OR MATCH" class to "TRLLINE."
16.Change the label of the "TRL THRU" class to "TRLTHRU."
17.Press .
Label the Calibration Kit
18.Press and create a label up to 8 characters long. For this example, enter "TRL KIT1" DONE.
19.To save the newly defined kit into nonvolatile memory, press:
Label the Classes

Calibrating for Increased Measurement Accuracy Calibrating for Non-Coaxial Devices (ES Analyzers Only)
Perform the TRL Calibration
Cal CAL KIT SELECT CAL KIT USER KIT RETURN RETURN CALIBRATE MENU TRL*/LRM* 2-PORT
J You could choose not to perform the isolation measurement by pressing
A poorly measured isolation class can actually degrade the overall measurement performance. If you are in doubt of the isolation measurement quality, you should omit the isolation portion of this procedure.
10.Connect a load to PORT 2, and press .
11.Connect the load to PORT 1, and press .
12.You may repeat any of the previous steps. There is no requirement to go in the order of steps. When the analyzer detects that you have made all the necessary measurements, the message line will show PRESS ’DONE’ IF FINISHED WITH CAL. Press
The message COMPUTING CAL COEFFICIENTS will appear, indicating that the analyzer is performing the numerical calculations of error coefficients.

Calibrating for Increased Measurement Accuracy Calibrating for Non-Coaxial Devices (ES Analyzers Only)
13.Connect the device under test. The device S-parameters are now being measured.


In order to use the LRM technique, the calibration standards characteristics must be entered into the analyzer’s user defined calibration kit.
The following steps show you how to define a calibration kit to utilize a set of LRM (LINE, REFLECT, MATCH) standards. This example LRM kit contains the following:
Modify the Standard Definitions
Preset Cal CAL KIT MODIFY DEFINE STANDARD
SHORT MODIFY STD DEFINITION SPECIFY OFFSET OFFSET DELAY 0
x1 STD OFFSET DONE STD DONE (DEFINED)
DEFINE STANDARD 4 x1 DELAY/THRU MODIFY STD DEFINITION
SPECIFY OFFSET OFFSET DELAY 0 x1 STD OFFSET DONE STD DONE (DEFINED)
DEFINE STANDARD 3 x1 LOAD

Calibrating for Increased Measurement Accuracy
Assign the Standards to the Various LRM Classes
Cal CAL KIT MODIFY SPECIFY CLASS MORE MORE TRL REFLECT
10.Since you previously designated standard #3 for the LINE/MATCH standard, press:
TRL LINE OR MATCH 3 x1
11.Since you previously designated standard #4 for the THRU/LINE standard, press:
TRL THRU 4 x1
12.To complete the specification of class assignments, press:
13.Press .
14.Change the label of the "TRL REFLECT" class to "LRMSHORT."
15.Change the label of the "TRL LINE OR MATCH" class to "LRMLOAD."
16.Change the label of the "TRL THRU" class to "LRMTHRU."
17.Press .
Label the Calibration Kit
18.Press and create a label up to 8 characters long. For this example, enter "LRM KIT1" and press .
DONE
19.To save the newly defined kit into nonvolatile memory, press:
LRM Error Correction · Label the Classes

Calibrating for Increased Measurement Accuracy LRM Error Correction
Perform the LRM Calibration
Cal CAL KIT SELECT CAL KIT USER KIT RETURN RETURN CALIBRATE MENU TRL*/LRM* 2-PORT
softkey.
SLIDING SLIDE IS SET
10.Connect the load to PORT 2 and press .
11.Repeat the previous LRM load measurement steps for PORT 2.
12.After the measurement is complete, press:
13.To measure the "ISOLATION" class, press:
J You could choose not to perform the isolation measurement by pressing

Calibrating for Increased Measurement Accuracy
To perform the best isolation measurements, you should reduce the system bandwidth or activate the averaging function.
A poorly measured isolation class can actually degrade the overall measurement performance. If you are in doubt of the isolation measurement quality, you should omit the isolation portion of this procedure.
14.You may repeat any of the previous steps. There is no requirement to go in the order of steps. When the analyzer detects that you have made all the necessary measurements, the message line will show PRESS ’DONE’ IF FINISHED WITH CAL. Press
The message COMPUTING CAL COEFFICIENTS will appear, indicating that the analyzer is performing the numerical calculations of error coefficients.
15.Connect the device under test. The device S-parameters are now being measured.
LRM Error Correction

Calibrating for Increased Measurement Accuracy Calibrating Using Electronic Calibration (ECal)
This section describes Electronic Calibration (ECal).
Use the following steps to perform the calibration.
J If you want to make a reflection measurement on PORT 1 (in the forward direction,
S11), leave the instrument default setting or press:
or on ET models:
J If you want to make a transmission measurement in the forward direction (S21),
press:
or on ET models:
J For ES analyzers, if you want to make a transmission measurement in the reverse
direction (S12), press:
J For ES analyzers, if you want to make a reflection measurement on PORT 2 (in the
reverse direction, S22), press:
Preset

Calibrating for Increased Measurement Accuracy Calibrating Using Electronic Calibration (ECal)
Figure 6-21 ECal Setup

If the cable is connected to the Test Set - I/O Interconnect connector, damage to the PC interface unit could occur.

Calibrating for Increased Measurement Accuracy Calibrating Using Electronic Calibration (ECal)
If the frequency span of the measurement that you set up earlier exceeds the span of a single ECal module, you need to use another ECal module whose frequency range allows the rest of the measurement span to be calibrated. The frequency range of the ECal modules is listed in the “General Information” chapter of the 85097A Electronic Calibration System User's Guide.
In addition to the standard ECal method, there are two options that you may want to use when performing the electronic calibration. They are:
Calibration using a manual thru is more accurate than calibrating using the thru internal to the ECal module. (A zero-length thru is created by connecting the two test port cables together.) The improved loss of the manual thru compared to the ECal module's thru, increases the accuracy of other error terms in the correction. The accuracy of the overall calibration is improved. Manual thrus can only be used with ECal modules having connectors of the opposite sex.

Calibrating for Increased Measurement Accuracy Calibrating Using Electronic Calibration (ECal)
During the isolation measurement portion of ECal, you are actually measuring instrument crosstalk. Typically, the data during this measurement is near the noise floor. (See also "Omitting Isolation Calibration" on page 6-4.)
When the crosstalk is near (or in) the noise floor, one way to reduce the noise is to turn on the isolation averaging. When the random noise of the instrument is averaged, its magnitude declines. As the energy of the trace is averaged, the displayed data becomes smoother. When the random noise is reduced, the network analyzer display shows the non-random crosstalk data.
To Select the Manual Thru Calibration Option
Cal ECal MENU CONFIGURE [ ]
MAN’L THRU on OFF
To Select the Isolation Calibration Option
Cal ECal MENU CONFIGURE [ ]
OMIT ISOLATION ON off
The isolation measurement is normally off. Therefore, the default setting for the
softkey is ON. When isolation is not omitted ( ), isolation standards are measured.
OMIT ISOLATION ON off
OMIT ISOLATION on OFF

Calibrating for Increased Measurement Accuracy Calibrating Using Electronic Calibration (ECal)
When ECal is first selected (or when you select module A or module B), there is a small initial delay so that the network analyzer can detect and download the calibration information from the internal memory of the ECal module.
— performs a measurement calibration for reflection-only. Measures of one-port devices or properly terminated two-port devices, at port 1 of an S-parameter test set. — performs a measurement calibration for reflection-only. Measures of one-port devices or properly terminated two-port devices, at port 2 of an S-parameter test set. — performs a complete calibration for measurement of all four S-parameters of a two-port device. This is the most accurate calibration for measurements of two-port devices. — performs an S11 and S21 enhanced response calibration (forward direction). Enhanced response generates a 1-port cal for S11 and an improved calibration over the response cal for S21. — performs an S22 and S12 enhanced response calibration (reverse direction). Enhanced response generates a 1-port cal for S22 and an improved calibration over the response cal for S12.
S22 1-PORT
FULL 2-PORT
S11/S21 ENH. RESP.
S22/S12 ENH. RESP.
— performs a reflection 1-port calibration. — performs a transmission (same as S21) and reflection (same as S11) enhanced response calibration (forward direction). Enhanced response generates a 1-port cal for reflection and an improved calibration over the response cal for transmission.
REFLECTION 1-PORT TRANS/REFL ENH. RESP.
Once the ECal calibration selection is made, the network analyzer performs the calibration.
Cal ECal MENU · MODULE A b

Calibrating for Increased Measurement Accuracy Calibrating Using Electronic Calibration (ECal)
Figure 6-22 Manual Thru Setup

Figure 6-23 Connect the Second ECal Module

CONTINUE ECal

Calibrating for Increased Measurement Accuracy Calibrating Using Electronic Calibration (ECal)
10.Save the calibration results by pressing .
Display information about the module by pressing .
Cal ECal MENU CONFIGURE [ ] MODULE INFO
A text window is displayed that contains the following information about the selected module.
1 PC interface unit port to which it is connected 6 Number of measurement points
2 Model number 7 Start (minimum) frequency
3 Serial number 8 Stop (maximum) frequency
4 Connector type 9 Suggested warmup time
5 Date of last certification 10 Warmup status (This is the remaining time on an internal analyzer timer. The timer simply counts down the recommended ECal module warm-up time. When the timer reaches zero, a message is displayed indicating that the ECal module is ready. The ECal module temperature or status is not read. The timer is reset when the analyzer is powered up or preset or when a new module is attached.)
CONTINUE ECal · Save/Recall

Calibrating for Increased Measurement Accuracy Calibrating Using Electronic Calibration (ECal)
The confidence check is a means of visually checking the quality of the calibration. The confidence check displays the currently measured data (DATA trace) and the factory-premeasured data (MEM trace) for the module’s confidence state. The confidence state, an independent reference standard, was not used for the calibration. For a good calibration measurement, there should be no significant difference between the traces (see Figure 6-24).
To Perform the Confidence Check
Cal ECal MENU CONFIGURE [ ]
Pressing the softkey toggles between the S11, S21, S12, and S22 (or for the ET models, TRANS (transmission) and REFL (reflection)) calibration traces. The confidence check will only display the S-parameter calibration data for which you calibrated. Uncalibrated S-parameter data traces are invalid. The memory trace still displays the module’s factory-premeasured trace.
Only one calibrated S-parameter
An enhanced response calibration
Only two calibrated S-parameters, one at a time
S11/S21 ENH. RESP.
S22/S12 ENH. RESP.
TRANS/REFL ENH. RESP.
A full 2-port calibration
All four calibrated S-parameters, one at a time
MODULE A b · For example, · A 1-port calibration

Calibrating for Increased Measurement Accuracy Calibrating Using Electronic Calibration (ECal)
Pressing the softkey toggles between the five trace-type display options. The confidence check can display the measured ECal results (DATA) and the premeasured calibration data (MEM) in following five ways.
The softkey is located in this menu for convenience in viewing the confidence check data. It acts the same as under the key.
AUTO SCALE Scale Ref
Figure 6-24 Confidence Check Display showing DATA&MEM Trace Type

![Page 361 of the Agilent Technologies Network Analyzer (20ES) user manual — REFL STD [ ]](/cdn/reader/15/150a289e983abebdc4bd6f0c3204b03f577bfd829d08c0cf5bb1353a14e103d7/images/page-361.webp?w=400)
Calibrating for Increased Measurement Accuracy Calibrating Using Electronic Calibration (ECal)
CAUTION The confidence check described in the previous section displays the ECal data of a single state. This confidence state is a calibrated standard not used during ECal. It is provided to give an independent assessment of the quality of a calibration. In the Ecal Service menu, you may also display each of the calibration standards which are used during an ECal calibration, along with the analyzer’s measurement of those standards.
You may notice a difference in measurement results when comparing ECal confidence state data and ECal standard state data. This result may be related to certain measurement errors with the network analyzer system, which add uncertainty to measurement results. More detailed information regarding measurement uncertainty is documented in the “Determining System Measurement Uncertainties” chapter of the network analyzer's reference guide. Additional information on improving your measurements can be found in Chapter 5 , “Optimizing Measurement Results.”
Using the ECal Service menu is not a standard part of the ECal procedure. It is a tool to allow you to identify problems in the calibration equipment, cables, connectors, or procedures.
The Confidence Check menu supports the comparison of the measured data versus the module's premeasured calibration data for the confidence state. The ECal Service menu supports the comparison of the measured data versus the module's premeasured calibration data for the remainder of the module states, in addition to the confidence state.
Access the ECal Service menu by pressing from the Confidence Check menu.
The ECal Service menu softkeys are:
ECal STD [ ]
Toggles the analyzer to show the data for the following calibration states:
Toggles between the available S11 and S22 reflection states listed in the softkey described above.
ECal STD [ ]
ECal SERVICE

Calibrating for Increased Measurement Accuracy Calibrating Using Electronic Calibration (ECal)
Toggles the analyzer to show the data for the following S-parameters:
Toggles through the following trace types:
Changes scale and reference values to bring the trace data in view on the display. The analyzer determines the smallest possible scale factor that will put all displayed data onto 80% of the vertical graticule. The reference value is chosen to put the trace in center screen.
Returns to the ECal Confidence Check menu.

Calibrating for Increased Measurement Accuracy Adapter Removal Using ECal (ES Analyzers Only)
A device under test (DUT) whose connectors cannot be connected directly to a test configuration is considered to be a noninsertable device. See Figure 6-25 Noninsertable devices can be caused because the DUT has:
Figure 6-25 Noninsertable Device
The adapter removal calibration technique may be used with the ECal calibration to accurately measure noninsertable devices. The following adapters (shown in Figure 6-26) are needed.

Calibrating for Increased Measurement Accuracy Adapter Removal Using ECal (ES Analyzers Only)
Figure 6-26 Adapters Needed
The following requirements must also be met:
For each port, a separate 2-port error correction needs to be performed to create two calibration sets. The adapter removal algorithm uses the resultant data from the two calibration sets and the nominal electrical length of the adapter to compute the adapter's actual S-parameters. This data is then used to generate a separate third calibration set in which the forward and reverse match and tracking terms are as if port 1 and port 2 could be connected. This is possible because the actual S-parameters of the adapter are measured with great accuracy, thus allowing the effects of the adapter to be completely removed when the third calibration set is generated.

Calibrating for Increased Measurement Accuracy Adapter Removal Using ECal (ES Analyzers Only)
Figure 6-27 Two-Port Cal Set 1

Cal ECal MENU MODULE A b

Calibrating for Increased Measurement Accuracy Adapter Removal Using ECal (ES Analyzers Only)
Figure 6-28 Two-Port Cal Set 2

10.Save the results to disk. Name the file "PORT2."
11.Determine the electrical delay of adapter A3. If you have adapter specifications that identify the electrical delay, you may use that information and continue with "Remove the Adapter" on page 6-74. If you do not know the delay of adapter A3, perform the “Determine the Electrical Delay” procedure on page 6-73.
Cal ECal MENU

Calibrating for Increased Measurement Accuracy Adapter Removal Using ECal (ES Analyzers Only)
This procedure determines the electrical delay of adapter A3 using a short.
This 1-port calibration can either be a manual calibration or an ECal.
Figure 6-29 Determining the Electrical Delay Setup
Format DELAY

Calibrating for Increased Measurement Accuracy Adapter Removal Using ECal (ES Analyzers Only)
When the two sets of error correction files have been created (now referred to as "calibration sets"), the A3 adapter may be removed.
Cal MORE ADAPTER REMOVAL
(This Help softkey provides a quick reference guide to using the adapter removal technique.)
ADAPTER DELAY
ADAPTER COAX
ADAPTER WAVEGUIDE
REMOVE ADAPTER
RECALL CAL SETS RECALL CAL PORT 1 RECALL CAL PORT 2
also displays the internal (or external if internal not used) disk file directory.
This process uses up an internal memory register. The calibration in this register is not the calibration created by adapter removal, rather it is a "scratch" calibration. You may wish to delete the register, or re-save the new calibration in this register as shown in the following step.
Save/Recall SELECT DISK INTERNAL MEMORY RETURN SAVE STATE

Calibrating for Increased Measurement Accuracy Adapter Removal Using ECal (ES Analyzers Only)
10.Connect the DUT to the network analyzer as shown in Figure 6-30 to perform calibrated measurements.
Figure 6-30 Calibrated Measurement
Since the effect of the adapter has been removed, it is easy to verify the accuracy of the technique by simply measuring the adapter itself. Because the adapter was used during the creation of the two calibration sets, and the technique removes its effects, measurement of the adapter itself should show the S-parameters.
If unexpected phase variations are observed, this indicates that the electrical delay of the adapter was not specified within a quarter wavelength over the frequency range of interest. To correct this, recall both cal sets, since the data was previously stored to disk, change the adapter delay, and press .

Calibrating for Increased Measurement Accuracy Adapter Removal Using ECal (ES Analyzers Only)


Operating Concepts Using This Chapter
This chapter provides conceptual information on how specific functions of the network analyzer operate. The following topics are discussed:

Network analyzers measure the reflection and transmission characteristics of devices and networks. A network analyzer test system consists of the following:
The analyzer applies a signal that is transmitted through the test device, or reflected from its input, and then compares it with the incident signal generated by the swept RF source. The signals are then applied to a receiver for measurement, signal processing, and display.
The vector network analyzer integrates a high resolution synthesized RF source, test set, and a dual channel three-input receiver to measure and display magnitude, phase, and group delay of transmitted and reflected power. With Option 010, the analyzer has the additional capability of transforming measured data from the frequency domain to the time domain. Other options are explained in “Options and Accessories” chapter of the reference guide. A simplified block diagram of the network analyzer system is shown in Figure 7-1. A detailed block diagram of the analyzer is provided in the service guide together with a theory of system operation.
Figure 7-1 Simplified Block Diagram of the Network Analyzer System
The analyzer's built-in synthesized source produces a swept RF signal or CW (continuous wave) signal in the following ranges:
Operating Concepts · System Operation

Operating Concepts System Operation
The RF output power is leveled by an internal ALC (automatic leveling control) circuit. To achieve frequency accuracy and phase measuring capability, the analyzer is phase locked to a highly stable crystal oscillator. For this purpose, a portion of the transmitted signal is routed to the R channel input of the receiver, where it is sampled by the phase detection loop and fed back to the source.
The Source Step Attenuator
The step attenuator contained in the source is used to adjust the power level to the test device without changing the level of the incident power in the reference path.
The ES model features a built-in test set that provides signal-separation devices and connections to the device under test. The signal separation devices are needed to separate the incident signal from the transmitted and reflected signals. The incident signal is applied to the R channel input through a jumper cable on the front panel. Meanwhile, the transmitted and reflected signals are internally routed from the test port couplers to the inputs of the A and B sampler/mixers in the receiver. Port 1 is connected to the A input and port 2 is connected to the B input. A transfer switch routes the source output power to either port 1 or 2, allowing measurements to be made in both the forward and reverse directions.
The ET model features a built-in test set that provides signal-separation and connections to the device under test. Part of the source output is coupled off and applied to the R input as the incident signal. The reflected signal is routed from the port 1 coupler to the input of the A sampler. The transmitted signal is routed from port 2 to the B sampler input.
The receiver block contains three sampler/mixers for the R, A, and B inputs. Instruments equipped with Option 400 contain an additional R sampler/mixer. The signals are sampled, and mixed to produce a 4 kHz IF (intermediate frequency). A multiplexer sequentially directs each of the three IF signals (four IF signals in Option 400) to the ADC (analog to digital converter) where it is converted from an analog to a digital signal. The signals are then measured and processed for viewing on the display. Both amplitude and phase information are measured simultaneously, regardless of what is displayed on the analyzer.
A microprocessor takes the raw data and performs all the required error correction, trace math, formatting, scaling, averaging, and marker operations, according to the instructions from the front panel or over GPIB. The formatted data is then displayed. The data processing sequence is described in “Processing” on page 7-5.
Measurements require calibration standards for vector accuracy enhancement (error-correction), and cables for interconnections. Model numbers and details of compatible power splitters, calibration kits, and cables are provided in “Options and Accessories” chapter of the reference guide.

The analyzer’s receiver converts the R, A, and B input signals into useful measurement information. This conversion occurs in two main steps:
The following paragraphs describe the sequence of math operations and the resulting data arrays as the information flows from the ADC to the display. They provide a good foundation for understanding most of the response functions, and the order in which they are performed.
Figure 7-2 is a data processing flow diagram that represents the flow of numerical data from IF detection to display. The data passes through several math operations, denoted in the figure by single line boxes. Most of these operations can be selected and controlled with the front panel response block menus. The data, stored in arrays along the way and denoted by double line boxes, are places in the flow path where data is accessible via GPIB.
Figure 7-2 Data Processing Flow Diagram
Operating Concepts · Processing

While only a single flow path is shown, two identical paths are available, corresponding to channel 1 and channel 2. Each channel also has an auxiliary channel for which the data is processed along with the primary channel’s data. Channel 3 is the auxiliary channel for channel 1, while channel 4 is the auxiliary channel for channel 2. When the channels are uncoupled, each channel is processed and controlled independently.
Data point definition: A "data point" or "point" is a single piece of data representing a measurement at a single stimulus value. Most data processing operations are performed point-by-point; some involve more than one point.
Sweep definition: A "sweep" is a series of consecutive data point measurements, taken over a sequence of stimulus values. A few data processing operations require that a full sweep of data is available. The number of points per sweep can be defined by the user. The units of the stimulus values (such as power, frequency, and time) can change, depending on the sweep mode, although this does not generally affect the data processing path.
The ADC (analog-to-digital converter) converts the R, A, and B inputs (already down-converted to a fixed low frequency IF) into digital words. (The AUX INPUT connector on the rear panel is a fourth input.) The ADC switches rapidly between these inputs, so they are converted nearly simultaneously.
This detection occurs in the digital filter, which performs the discrete Fourier transform (DFT) on the digital words. The samples are converted into complex number pairs (real plus imaginary, R+jX). The complex numbers represent both the magnitude and phase of the IF signal. If the AUX INPUT is selected, the imaginary part of the pair is set to zero. The DFT filter shape can be altered by changing the IF bandwidth, which is a highly effective technique for noise reduction.
These calculations are performed if the selected measurement is a ratio of two inputs (for example, A/R or B/R). This is a complex divide operation. If the selected measurement is absolute (such as A or B), no calculations are performed. The R, A, and B values are also split into channel data at this point.
The next digital processing technique used is sampler/IF correction. This process digitally corrects for frequency response errors (both magnitude and phase, primarily sampler rolloff) in the analog down-conversion path.
Averaging is another noise reduction technique. This calculation involves taking the complex exponential average of several consecutive sweeps. This technique cannot be used with single-input measurements.

These data arrays store the results of all the preceding data processing operations. (Up to this point, all processing is performed real-time with the sweep by the IF processor. The remaining operations are not necessarily synchronized with the sweep, and are performed by the main processor.) When full 2-port error correction is on, the raw arrays contain all four S-parameter measurements required for accuracy enhancement. When the channels are uncoupled ( ), there may be as many as eight raw arrays. These arrays are directly accessible via GPIB. Notice that the numbers here are still complex pairs.
Raw arrays contain the pre-raw data which has sampler and attenuator offset applied.
Vector Error-correction (Accuracy Enhancement)
Error-correction is performed next, if a measurement calibration has been performed and correction is activated. Error-correction removes repeatable systematic errors (stored in the error coefficient arrays) from the raw arrays. This can vary from simple vector normalization to full 12-term error-correction.
The results of error-correction are stored in the data arrays as complex number pairs. These are subsequently used whenever correction is on, and are accessible via GPIB.
If the data-to-memory operation is performed, the data arrays are copied into the memory arrays.
This operation selects either the data array, memory array, or both to continue flowing through the data processing path. In addition, the complex ratio of the two (data/memory) or the difference (data−memory) can also be selected. If memory is displayed, the data from the memory arrays goes through exactly the same processing flow path as the data from the data arrays.
Gating (Option 010 Only)
This digital filtering operation is associated with time domain transformation. Its purpose is to mathematically remove unwanted responses isolated in time. In the time domain, this can be viewed as a time-selective bandpass or bandstop filter. (If both data and memory are displayed, gating is applied to the memory trace only if gating was on when data was stored into memory.)
The Electrical Delay Block
This block involves adding or subtracting phase in proportion to frequency. This is equivalent to "line-stretching" or artificially moving the measurement reference plane. This block also includes the effects of port extensions as well as electrical delay.
This converts the measured S-parameter data to the equivalent complex impedance (Z) or admittance (Y) values, or to inverse S-parameters (1/S).

Transform (Option 010 Only)
This transform converts frequency domain information into the time domain when it is activated. The results resemble time domain reflectometry (TDR) or impulse-response measurements. The transform uses the chirp-Z inverse fast Fourier transform (FFT) algorithm to accomplish the conversion. The windowing operation, if enabled, is performed on the frequency domain data just before the transform. (A special transform mode is available to "demodulate" CW sweep data, with time as the stimulus parameter, and display spectral information with frequency as the stimulus parameter.)
This operation converts the complex number pairs into a scalar representation for display, according to the selected format. This includes group delay calculations. These formats are often easier to interpret than the complex number representation. (Polar and Smith chart formats are not affected by the scalar formatting.) It is impossible to recover the complex data after formatting, as shown in Figure 7-2.
This noise reduction technique smoothes noise on the trace. Smoothing is also used to set the aperture for group delay measurements.
When smoothing is on, each point in a sweep is replaced by the moving average value of several adjacent (formatted) points. The number of points included depends on the smoothing aperture, which can be selected by the user. The effect is similar to video filtering. If data and memory are displayed, smoothing is performed on the memory trace only if smoothing was on when data was stored into memory.
The data processing results are now stored in the format arrays. Notice that the marker values and marker functions are all derived from the format arrays in Figure 7-2. Limit testing is also performed on the formatted data. The format arrays are accessible via GPIB.
These operations prepare the formatted data for display. This is where the reference line position, reference line value, and scale calculations are performed, as appropriate to the format.
The display memory stores the display image for presentation on the analyzer. The information stored includes graticules, annotation, and softkey labels. If user display graphics are written, these are also stored in display memory. When a print or plot is made, the information is taken from display memory.
The display is updated frequently and synchronously with the data processing operations.

The built-in synthesized source contains a programmable step attenuator that allows you to directly and accurately set power levels in twelve different power ranges. Each range has a total span of 20 dB (15 dB, 8722ET/ES). The twelve ranges cover the instrument’s full operating range. In addition, some amount of overrange and underrange is permitted beyond the stated limits. Performance outside of the range limits varies from one analyzer to another and is not specified. A power range can be selected either manually or automatically.
If you select , you can enter any power level within the total operating range of the instrument and the source attenuator will automatically switch to the corresponding range.
Each range overlaps its adjacent ranges by 15 dB, therefore, certain power levels are designated to cause the attenuator to switch to the next range so that optimum (leveled) performance is maintained. These transition points exist at −10 dB from the top of a range and at +5 dB from the bottom of a range. This leaves 10 dB of operating range. By turning the RPG knob with being the active function, you can hear the attenuator switch as these transitions occur. See the specifications and characteristics chapter in the reference guide for the analyzer’s power limits.
If you select , you must first enter the power ranges menu and manually select the power range that corresponds to the power level you want to use. This is accomplished by pressing the softkey and then selecting one of the 12 available ranges. The active power will always be reset to be within this range. In this mode, you will be able to select power levels above or below the range limits. The factory limits on setting power level are +6 dB above the maximum range power and −12 dB below the minimum range power. Typical performance for power overrange is 2 dB above the maximum (more at lower frequencies and less at higher frequencies) and typical under range is 12 dB. An under-range condition may cause a phase lock lost message to be displayed.
POWER RANGES
When a calibration is active, the power range selection is switched from auto to manual mode, and PRm appears on the display. This feature is necessary to maintain accuracy once a measurement calibration is activated.
![Page 380 of the Agilent Technologies Network Analyzer (20ES) user manual — PORT PWR [COUPLED]](/cdn/reader/15/150a289e983abebdc4bd6f0c3204b03f577bfd829d08c0cf5bb1353a14e103d7/images/page-380.webp?w=400)
Operating Concepts Output Power
If you decide to switch power ranges, the calibration accuracy is degraded and accuracy is no longer specified. However, the analyzer leaves the correction on.
The annotation C∆ will be displayed whenever you change the power after calibration.
There are two methods you can use to couple and uncouple power levels with the analyzer:
By uncoupling the channel powers, you effectively have two separate sources. Uncoupling the test ports allows you to have different power levels on each port.
toggles between coupled and uncoupled channel power. With the channel power coupled, the power levels are the same on each channel. With the channel power uncoupled, you can set different power levels for each channel. For the channel power to be uncoupled, the other channel stimulus functions must also be uncoupled ( ).
COUPLED CH OFF
toggles between coupled and uncoupled test ports. With the test ports coupled, the power level is the same at each port. With the ports uncoupled, you can set a different power level at each port. This can be useful, for example, if you want to simultaneously perform a gain and reverse isolation measurement on a high-gain amplifier using the dual channel mode to display the results. In this case, you would want the power in the forward direction (S21) much lower than the power in the reverse direction (S12).
Channel coupling · Test port coupling
![Page 381 of the Agilent Technologies Network Analyzer (20ES) user manual — SWEEP TIME [AUTO]](/cdn/reader/15/150a289e983abebdc4bd6f0c3204b03f577bfd829d08c0cf5bb1353a14e103d7/images/page-381.webp?w=400)
The softkey selects sweep time as the active entry and shows whether the automatic or manual mode is active. The following explains the difference between automatic and manual sweep time:
Sweep time refers only to the time that the instrument is sweeping and taking data, and does not include the time required for internal processing of the data, retrace time, or band switching time. A sweep speed indicator ↑ is displayed on the trace for sweep times longer than 1.0 second. For sweep times equal to or faster than 1.0 second, the ↑ indicator appears in the status notations area at the left of the analyzer's display.
When this mode is active, the softkey label reads . This mode is engaged whenever you enter a sweep time greater than zero. This mode allows you to select a fixed sweep time. If you change the measurement parameters such that the current sweep time is no longer possible, the analyzer will automatically increase to the next fastest sweep time possible. If the measurement parameters are changed such that a faster sweep time is possible, the analyzer will not alter the sweep time while in this mode.
Auto Sweep Time Mode
When this mode is active, the softkey label reads . This mode is engaged whenever you enter as a sweep time. Auto sweep time continuously maintains the fastest sweep time possible with the selected measurement parameters.
Minimum Sweep Time
The minimum sweep time is dependent on the following measurement parameters:
Operating Concepts · Sweep Time

Operating Concepts Sweep Time
In addition to the these parameters, the actual cycle time of the analyzer is also dependent on the following measurement parameters:
Refer to the specifications and characteristics chapters of the reference guide to see the minimum cycle time values for specific measurement parameters.

Operating Concepts Source Attenuator Switch Protection
The programmable step attenuator of the source can be switched between port 1 and port 2 when the test port power is uncoupled, or between channel 1 and channel 2 when the channel power is uncoupled. To avoid premature wear of the attenuator, measurement configurations requiring continuous switching between different power ranges are not allowed.
For example, continuous switching would be required if channels 1 and 2 of the analyzer are decoupled, power levels in two different ranges are selected for each channel, and dual channel display is engaged. To prevent continuous switching between the two power ranges, the analyzer automatically engages the test set hold mode after measuring both channels once. The active channel continues to be updated each sweep while the inactive channel is placed in the hold mode. (The status annotation tsH appears on the left side of the display.) If averaging is on, the test set hold mode does not engage until the specified number of sweeps is completed. The and softkeys can override this protection feature.
Allowing Repetitive Switching of the Attenuator
The and softkeys allow measurements which demand repetitive switching of the step attenuator. Use these softkeys with caution; repetitive switching can cause premature wearing of the attenuator.

Operating Concepts Channel Stimulus Coupling
toggles the channel coupling of stimulus values. With (the preset condition), both channels have the same stimulus values. (The inactive channel takes on the stimulus values of the active channel.)
COUPLED CH on OFF COUPLED CH ON
In the stimulus coupled mode, the following parameters are coupled:
Coupling of stimulus values for the two channels is independent of in the display menu and in the marker mode menu. activates an alternate sweep function when dual channel display is on. In this mode, the analyzer alternates between the two sets of stimulus values and displays the measurement data of both channels.
![Page 385 of the Agilent Technologies Network Analyzer (20ES) user manual — LIST FREQ [STEPPED]](/cdn/reader/15/150a289e983abebdc4bd6f0c3204b03f577bfd829d08c0cf5bb1353a14e103d7/images/page-385.webp?w=400)
The following sweep types will function with the interpolated error-correction feature (described in “Interpolated Error Correction” on page 6-8):
The following sweep types will not function with the interpolated error correction feature:
The softkey activates a linear frequency sweep that is displayed on a standard graticule with ten equal horizontal divisions. This is the preset default sweep type.
For a linear sweep, sweep time is combined with the channel's frequency span to compute a source sweep rate:
sweep rate = (frequency span) / (sweep time)
Since the sweep time may be affected by various factors, the equation provided here is merely an indication of the ideal (fastest) sweep rate. If the user-specified sweep time is greater than 15 ms times the number of points, the sweep changes from a continuous ramp sweep to a stepped CW sweep. Also, for 10 Hz or 30 Hz IF bandwidths, the sweep is automatically converted to a stepped CW sweep.
In the linear frequency sweep mode, it is possible, with Option 010, to transform the data for time domain measurements using the inverse Fourier transform technique.
The softkey activates a logarithmic frequency sweep mode. The source is stepped in logarithmic increments and the data is displayed on a logarithmic graticule. This is slower than a continuous sweep with the same number of points, and the entered sweep time may therefore be changed automatically. For frequency spans of less than two octaves, the sweep type automatically reverts to linear sweep.
Stepped List Frequency Sweep (Hz)
The softkey activates a stepped list frequency sweep, one of two list frequency sweep modes. The stepped list mode allows the analyzer to sweep a list of arbitrary frequency points. This list is defined and modified using the edit list menu and the edit subsweep menu. Up to 30 frequency subsweeps (called "segments") of several different types can be specified, for a maximum total of 1601 points.
Operating Concepts · Sweep Types

Operating Concepts Sweep Types
One list is common to both channels. Once a frequency list has been defined and a measurement calibration performed on the full frequency list, one or all of the frequency segments can be measured and displayed without loss of calibration.
When the key is pressed, the network analyzer sorts all the defined frequency segments into CW points in order of increasing frequency. It then measures each point and displays a single trace that is a composite of all data taken. If duplicate frequencies exist, the analyzer makes multiple measurements on identical points to maintain the specified number of points for each subsweep. Since the frequency points may not be distributed evenly across the display, the display resolution may be uneven, and more compressed in some parts of the trace than in others. However, the stimulus and response readings of the markers are always accurate. Because the list frequency sweep is a stepped CW sweep, the sweep time is slower than for a continuous sweep with the same number of points.
The softkey provides access to the segment menu, which allows you to select any single segment ( ) in the frequency list or all of the segments ( ) in the frequency list.
SINGLE SEG SWEEP ALL SEGS SWEEP
See the following information on how to enter or modify the list frequencies. If no list has been entered, the message CAUTION: LIST TABLE EMPTY is displayed. A tabular printout of the frequency list data can be obtained using the function in the copy menu.
Stepped Edit List Menu
The softkey within the sweep type menu provides access to the edit list menu.
This menu is used to edit the list of frequency segments (subsweeps) defined with the edit subsweep menu, described next. Up to 30 frequency subsweeps can be specified, for a maximum of 1601 points. The segments do not have to be entered in any particular order: the analyzer automatically sorts them and shows them on the display in increasing order of start frequency. This menu determines which entry on the list is to be modified, while the edit subsweep menu is used to make changes in the frequency or number of points of the selected entry.
Stepped Edit Subsweep Menu
Using the or softkey within the edit list menu will display the edit subsweep menu. This menu lets you select measurement frequencies arbitrarily. Using this menu it is possible to define the exact frequencies to be measured on a point-by-point basis. For example, the sweep could include 100 points in a narrow passband, 100 points across a broad stop band, and 50 points across the third harmonic response. The total sweep is defined with a list of subsweeps.
Segment Menu

The frequency subsweeps, or segments, can be defined in any of the following terms:
The subsweeps can overlap, and do not have to be entered in any particular order. The analyzer sorts the segments automatically and lists them on the display in order of increasing start frequency, even if they are entered in center/span format. If duplicate frequencies exist, the analyzer makes multiple measurements on identical points to maintain the specified number of points for each subsweep. The data is shown on the display as a single trace that is a composite of all data taken. The trace may appear uneven because of the distribution of the data points, but the frequency scale is linear across the total range.
Once the list frequencies have been defined or modified, the list frequency sweep mode can be selected with the softkey in the sweep type menu. The frequency list parameters can also be saved with an instrument state.
Swept List Frequency Sweep (Hz)
The softkey activates a swept list frequency sweep, one of two list frequency sweep modes. The swept list mode allows the analyzer to sweep a list of arbitrary frequency points which are defined and modified in a way similar to the stepped list mode.
However, this mode takes data while sweeping through the defined frequency points, increasing throughput by up to 6 times over a stepped sweep. In addition, this mode allows the test port power and IF bandwidth to be set independently for each segment that is defined. The only restriction is that you cannot specify overlapping frequency segments.
Similar to stepped list mode, the softkey also provides access to the segment menu. However, swept list mode expands the way segments can be defined.Refer to the following information on how to enter or modify the list segments.
Swept Edit List Menu
The softkey within the sweep type menu provides access to the edit list menu. The function of this menu is the same as in the stepped list mode.
Swept Edit Subsweep Menu
Using the or softkey within the edit list menu will display the edit subsweep menu. This menu lets you select measurement frequencies arbitrarily. Using this menu it is possible to define the exact frequencies to be measured on a point-by-point basis at specific power levels and IF bandwidth settings. The total sweep is defined with a list of subsweeps.
Operating Concepts · Sweep Types
![Page 388 of the Agilent Technologies Network Analyzer (20ES) user manual — LIST FREQ [SWEPT]](/cdn/reader/15/150a289e983abebdc4bd6f0c3204b03f577bfd829d08c0cf5bb1353a14e103d7/images/page-388.webp?w=400)
Operating Concepts Sweep Types
The frequency subsweeps, or segments, can be defined in any of the following terms:
See “Setting Segment Power” and “Setting Segment IF Bandwidth” on page 7-19 for information on how to set the segment power and IF bandwidth.
The subsweeps may be entered in any particular order but they cannot overlap. The analyzer sorts the segments automatically and lists them on the display in order of increasing start frequency, even if they are entered in center/span format. The data is shown on the display as a single trace that is a composite of all data taken. The trace may appear uneven because of the distribution of the data points, but the frequency scale is linear across the total range.
Once the list frequencies have been defined or modified, the list frequency sweep mode can be selected with the softkey in the sweep type menu. The frequency list parameters can also be saved with an instrument state.
To enable the function, you must first select in the edit subsweep menu. List power is off by default and the asterisks that appear in the "power" column of the list table indicate that power for the sweep is being set by the normal analyzer power controls.
SEGMENT POWER LIST POWER ON off
The power settings for all segments are restricted to a single power range. This prevents the attenuator from switching to different settings mid-sweep. Select the power range and then edit the list table to specify the segment powers. If the power range is selected after the list has been defined, the list settings may be affected.
When analyzer port power is uncoupled, the segment power level can be set independently for each port. To do this, you must first select a measurement parameter to activate the port whose power you want to set. For example, select S11 to set port 1 power, or S22 to set port 2 power. (Notice that the list mode table will only display the currently selected port in the table. This is due to restricted display space.)
When analyzer port power is uncoupled, the softkey can also be set independently for each port. For example, you may choose to set for forward measurements and for reverse measurements. In this case, the power would be set according to values in the list when measuring the forward parameters. When measuring the reverse parameters, the power would be set according to the normal analyzer power controls.
LIST POWER ON off
LIST POWER ON off LIST POWER on OFF
Setting Segment Power

Setting Segment IF Bandwidth
To enable the function, you must first select in the edit subsweep menu. List IF bandwidth is off by default and the asterisks that appear in the "IFBW" column of the list table indicate that the IF bandwidth for the sweep is being set by the normal analyzer controls.
SEGMENT IF BW LIST IF BW ON off
Narrow IF bandwidths require more data samples per point and thus slow down the measurement time. Selectable IF bandwidths can increase the throughput of the measurement by allowing you to specify narrow bandwidths only where needed.
The softkey turns on a power sweep mode that is used to characterize power-sensitive circuits. In this mode, power is swept at a single frequency, from a start power value to a stop power value, selected using the and keys and the entry block. This feature is convenient for such measurements as gain compression or AGC (automatic gain control) slope. To set the frequency of the power sweep, use in the stimulus menu.
CW FREQ
The span of the swept power is limited to being equal to or within one of the eight pre-defined power ranges. The attenuator will not switch to a different power range while in the power sweep mode. Therefore, when performing a power sweep, power range selection will automatically switch to the manual mode.
In power sweep, the entered sweep time may be automatically changed if it is less than the minimum required for the current configuration (number of points, IF bandwidth, averaging, etc.).
The softkey turns on a sweep mode similar to an oscilloscope. The analyzer is set to a single frequency, and the data is displayed versus time. The frequency of the CW time sweep is set with in the stimulus menu. In this sweep mode, the data is continuously sampled at precise, uniform time intervals determined by the sweep time and the number of points minus 1. The entered sweep time may be automatically changed if it is less than the minimum required for the current instrument configuration.
CW FREQ
In time domain using Option 010, the CW time mode data is translated to frequency domain, and the x-axis becomes frequency. In this mode, the instrument can be used as a spectrum analyzer to measure signal purity, or for low frequency (
In addition to the previous sweep types, there are also two different sweep modes. These can be accessed through the correction menu by pressing or .
Cal MORE ALTERNATE A and B CHOP A and B
Operating Concepts · Sweep Types · Start Stop

The key accesses the S-parameter menu which contains softkeys that can be used to select the parameters or inputs that define the type of measurement being performed.
S-parameters (scattering parameters) are a convention used to characterize the way a device modifies signal flow. A brief explanation of the S-parameters of a two-port device is provided, however, for additional details, refer to Application Notes 95-1 and 154.
S-parameters are always a ratio of two complex (magnitude and phase) quantities. S-parameter notation identifies these quantities using the numbering convention:
where the first number (out) refers to the test-device port where the signal is emerging and the second number (in) is the test-device port where the signal is incident. For example, the S-parameter S21 identifies the measurement as the complex ratio of the signal emerging at the test device’s port 2 to the signal incident at the test device’s port 1.
Figure 7-3 is a representation of the S-parameters of a two-port device, together with an equivalent flowgraph. In the illustration, "a" represents the signal entering the device and "b" represents the signal emerging. Note that a and b are not related to the A and B input ports on the analyzer.
Operating Concepts S-Parameters · Meas · S out in

Figure 7-3 S-Parameters of a Two-Port Device
S-parameters are exactly equivalent to these more common description terms, requiring only that the measurements be taken with all test device ports properly terminated.
S-Parameter Definition Test Set Description Direction
S11 b1/a1 a2 = 0 Input reflection coefficient FWD
S21 b2/a1 a2 = 0 Forward gain FWD
S12 b1/a2 a1 = 0 Reverse Gain REV
S22 b2/a2 a1 = 0 Output reflection coefficient REV
Operating Concepts · S-Parameters

The S-parameter menu allows you to define the input ports and test set direction for S-parameter measurements. The analyzer automatically switches the direction of the measurement according to the selections you made in this menu. Therefore, the analyzer can measure all four S-parameters with a single connection. The S-parameter being measured is labeled at the top left corner of the display.
The S-parameter menu of the ES model analyzer contains the following softkeys:
ANALOG IN Aux Input
The S-parameter menu of the ET model analyzer contains the following softkeys:
TRANSMISSN
ANALOG IN Aux Input
This menu allows you to monitor voltage and frequency nodes, using the analog bus and internal counter. For more information, refer to the "Service Menus and Error Messages" chapter in the service guide.
This menu converts the measured reflection or transmission data to the equivalent complex impedance (Z) or admittance (Y) values. This is not the same as a two-port Y or Z parameter conversion, as only the measured parameter is used in the equations. Two simple one-port conversions are available, depending on the measurement configuration.
An S11 or S22 trace measured as reflection can be converted to equivalent parallel impedance or admittance using the model and equations shown in Figure 7-4.

Figure 7-4 Reflection Impedance and Admittance Conversions
In a transmission measurement, the data can be converted to its equivalent series impedance or admittance using the model and equations shown in Figure 7-5.
Figure 7-5 Transmission Impedance and Admittance Conversions
This menu allows you to define the input ports for power ratio measurements, or a single input for magnitude only measurements of absolute power. You cannot use single inputs for phase or group delay measurements, or any measurements with averaging activated.
Operating Concepts · S-Parameters · Input Ports Menu

Operating Concepts Analyzer Display Formats
The key accesses the format menu. This menu allows you to select the appropriate display format for the measured data.
The analyzer automatically changes the units of measurement to correspond with the displayed format. Special marker menus are available for the polar and Smith formats, each providing several different marker types for readout of values.
The selected display format of a particular S-parameter or input is assigned to that parameter. Thus if different S-parameters are measured, even if only one channel is used, each parameter is shown in its selected format each time it is displayed.
The following illustrations show a reflection measurement of a bandpass filter displayed in each of the available formats.
The softkey displays the log magnitude format. This is the standard Cartesian format used to display magnitude-only measurements of insertion loss, return loss, or absolute power in dB versus frequency. The bandpass filter reflection data in a log magnitude format is illustrated in Figure 7-6.
Figure 7-6 Log Magnitude Format

The softkey displays a Cartesian format of the phase portion of the data, measured in degrees. This format displays the phase shift versus frequency. The phase response of the same filter in a phase-only format is illustrated in Figure 7-7.
Format

Operating Concepts Analyzer Display Formats
Figure 7-7 Phase Format

The softkey selects the group delay format, with marker values given in seconds. The bandpass filter response formatted as group delay is shown in Figure 7-8. Group delay principles are described in the next few pages.
Figure 7-8 Group Delay Format


Operating Concepts Analyzer Display Formats
The softkey displays a Smith chart format. Refer to Figure 7-9. This is used in reflection measurements to provide a readout of the data in terms of impedance. The intersecting dotted lines on the Smith chart represent constant resistance and constant reactance values, normalized to the characteristic impedance, Z0, of the system. Reactance values in the upper half of the Smith chart circle are positive (inductive) reactance, and those in the lower half of the circle are negative (capacitive) reactance. The default marker readout is in ohms (Ω) to measure resistance and reactance (R+jX). Additional marker types are available in the Smith marker menu.
The Smith chart is most easily understood with a full scale value of 1.0. If the scale per division is less than 0.2, the format switches automatically to polar.
If the characteristic impedance of the system is not 50 ohms, modify the impedance value recognized by the analyzer by pressing (the impedance value) .
Cal MORE SET Z0 x1
An inverted Smith chart format for admittance measurements is also available. This is shown in Figure 7-9. Access this by selecting in the format menu, and pressing . The Smith chart is inverted and marker values are read out in siemens (S) to measure conductance and susceptance (G+jB).
Marker Fctn MKR MODE MENU SMITH MKR MENU G+jB MKR
Figure 7-9 Standard and Inverse Smith Chart Formats


Operating Concepts Analyzer Display Formats
The softkey displays a polar format as shown in Figure 7-10. Each point on the polar format corresponds to a particular value of both magnitude and phase. Quantities are read vectorally: the magnitude at any point is determined by its displacement from the center (which has zero value), and the phase by the angle counterclockwise from the positive x-axis. Magnitude is scaled in a linear fashion, with the value of the outer circle usually set to a ratio value of 1. Since there is no frequency axis, frequency information is read from the markers.
The default marker readout for the polar format is in linear magnitude and phase. A log magnitude marker and a real/imaginary marker are available in the polar marker menu.
Figure 7-10 Polar Format

The softkey displays the linear magnitude format as shown in Figure 7-11. This is a Cartesian format used for unitless measurements such as reflection coefficient magnitude ρ or transmission coefficient magnitude τ, and for linear measurement units. It is used for display of conversion parameters and time domain transform data.

Operating Concepts Analyzer Display Formats
Figure 7-11 Linear Magnitude Format

The softkey reformats a reflection measurement into its equivalent SWR (standing wave ratio) value. See Figure 7-12. SWR is equivalent to (1 + ρ)/(1 − ρ), where ρ is the reflection coefficient. Note that the results are valid only for reflection measurements. If the SWR format is used for measurements of S21 or S12 the results are not valid.
Figure 7-12 Typical SWR Display


Operating Concepts Analyzer Display Formats
The softkey displays only the real (resistive) portion of the measured data on a Cartesian format. See Figure 7-13. This is similar to the linear magnitude format, but can show both positive and negative values. It is primarily used for analyzing responses in the time domain, and also to display an auxiliary input voltage signal for service purposes.
Figure 7-13 Real Format

The softkey displays only the imaginary (reactive) portion of the measured data on a Cartesian format. This format is similar to the real format except that reactance data is displayed on the trace instead of resistive data.
Group Delay Principles
For many networks, the amount of insertion phase is not as important as the linearity of the phase shift over a range of frequencies. The analyzer can measure this linearity and express it in two different ways: directly, as deviation from linear phase, or as group delay, a derived value.
Group delay is the measurement of signal transmission time through a test device. It is defined as the derivative of the phase characteristic with respect to frequency. Since the derivative is basically the instantaneous slope (or rate of change of phase with respect to frequency), a perfectly linear phase shift results in a constant slope, and therefore a constant group delay. See Figure 7-14.

Operating Concepts Analyzer Display Formats
Figure 7-14 Constant Group Delay

Note, however, that the phase characteristic typically consists of both linear and higher order (deviations from linear) components. The linear component can be attributed to the electrical length of the test device, and represents the average signal transit time. The higher order components are interpreted as variations in transit time for different frequencies, and represent a source of signal distortion. See Figure 7-15.
Figure 7-15 Higher Order Phase Shift

The analyzer computes group delay from the phase slope. Phase data is used to find the phase change, ∆ Φ over a specified frequency aperture, ∆ f, to obtain an approximation for the rate of change of phase with frequency. Refer to Figure 7-16. This value, (τ)g, represents the group delay in seconds assuming linear phase change over ∆f. It is important that ∆ Φ be ≤ 180°, or errors will result in the group delay data. These errors can be significant for long delay devices. You can verify that ∆ Φ is ≤ 180° by increasing the number of points or narrowing the frequency span (or both) until the group delay data no longer changes.

Operating Concepts Analyzer Display Formats
Figure 7-16 Rate of Phase Change Versus Frequency

When deviations from linear phase are present, changing the frequency step can result in different values for group delay. Note that in this case the computed slope varies as the aperture ∆f is increased. See Figure 7-17. A wider aperture results in loss of the fine grain variations in group delay. This loss of detail is the reason that in any comparison of group delay data, it is important to know the aperture that was used to make the measurement.
Figure 7-17 Variations in Frequency Aperture

In determining the group delay aperture, there is a trade-off between resolution of fine detail and the effects of noise. Noise can be reduced by increasing the aperture, but this will tend to smooth out the fine detail. More detail will become visible as the aperture is decreased, but the noise will also increase, possibly to the point of obscuring the detail. A good practice is to use a smaller aperture to assure that small variations are not missed, then increase the aperture to smooth the trace.

Operating Concepts Analyzer Display Formats
The default group delay aperture is the frequency span divided by the number of points across the display. To set the aperture to a different value, turn on smoothing in the average menu, and vary the smoothing aperture. The aperture can be varied up to 20% of the span swept.
Group delay measurements can be made on linear frequency, log frequency, or list frequency sweep types (not in CW or power sweep). Group delay aperture varies depending on the frequency spacing and point density, therefore the aperture is not constant in log and list frequency sweep modes. In list frequency mode, extra frequency points can be defined to ensure the desired aperture.
To obtain a readout of aperture values at different points on the trace, turn on a marker. Then press . Smoothing aperture becomes the active function, and as the aperture is varied its value in Hz is displayed beneath the active entry area.
Avg SMOOTHING APERTURE

The softkey adjusts the electrical delay to balance the phase of the test device. This softkey must be used in conjunction with or (with cut-off frequency) in order to identify which type of transmission line the delay is being added to. These softkeys can be accessed by pressing the key.
COAXIAL DELAY WAVEGUIDE DELAY
Electrical delay simulates a variable length lossless transmission line, which can be added to or removed from a receiver input to compensate for interconnecting cables, etc. This function is similar to the mechanical or analog "line stretchers" of other network analyzers. Delay is annotated in units of time with secondary labeling in distance for the current velocity factor.
With this feature, and with (refer to “Setting the Electrical Delay” on page 1-37), an equivalent length of air-filled, lossless transmission line is added or subtracted according to the following formula:
Length meters ( ) Φ
Freq MHz ( ) 1.20083 × ( ) =
Once the linear portion of the test device's phase has been removed, the equivalent length of the lossless, transmission line can be read out in the active marker area. If the average relative permittivity (εr) of the test device is known over the frequency span, the length calculation can be adjusted to indicate the actual length of the test device more closely. This can be done by entering the relative velocity factor for the test device using the calibrate more menu. The relative velocity factor for a given dielectric can be calculated by:
Velocity Factor 1 εr =
assuming a relative permeability of 1.
Operating Concepts · Electrical Delay · Scale Ref

Operating Concepts Noise Reduction Techniques
The key is used to access three different noise reduction techniques: sweep-to-sweep averaging, display smoothing, and variable IF bandwidth. All of these can be used simultaneously. Averaging and smoothing can be set independently for each channel, and the IF bandwidth can be set independently if the stimulus is uncoupled.
Averaging computes each data point based on an exponential average of consecutive sweeps weighted by a user-specified averaging factor. Each new sweep is averaged into the trace until the total number of sweeps is equal to the averaging factor, for a fully averaged trace. Each point on the trace is the vector sum of the current trace data and the data from the previous sweep. A high averaging factor gives the best signal-to-noise ratio, but slows the trace update time. Doubling the averaging factor reduces the noise by 3 dB. Averaging is used for ratioed measurements: if it is attempted for a single-input measurement (e.g. A or B), the message CAUTION: AVERAGING INVALID ON NON-RATIO MEASURE is displayed. The effect of averaging on a log magnitude format trace is shown in Figure 7-18.
Figure 7-18 Effect of Averaging on a Trace

Avg

Operating Concepts Noise Reduction Techniques
Smoothing (similar to video filtering) averages the formatted active channel data over a portion of the displayed trace. Smoothing computes each displayed data point based on one sweep only, using a moving average of several adjacent data points for the current sweep. The smoothing aperture is a percent of the swept stimulus span, up to a maximum of 20%.
Rather than lowering the noise floor, smoothing finds the mid-value of the data. Use it to reduce relatively small peak-to-peak noise values on broadband measured data. Use a sufficiently high number of display points to avoid misleading results. Do not use smoothing for measurements of high resonance devices or other devices with wide trace variations, as it will introduce errors into the measurement.
Smoothing is used with Cartesian and polar display formats. It is also the primary way to control the group delay aperture, given a fixed frequency span. Refer to “Group Delay Principles” on page 7-29. In polar display format, large phase shifts over the smoothing aperture will cause shifts in amplitude, since a vector average is being computed. The effect of smoothing on a log magnitude format trace is illustrated in Figure 7-19.
Figure 7-19 Effect of Smoothing on a Trace

IF bandwidth reduction lowers the noise floor by digitally reducing the receiver input bandwidth. It works in all ratio and non-ratio modes. It has an advantage over averaging as it reliably filters out unwanted responses such as spurs, odd harmonics, higher frequency spectral noise, and line-related noise. Sweep-to-sweep averaging, however, is better at filtering out very low frequency noise. A tenfold reduction in IF bandwidth lowers the measurement noise floor by about 10 dB. Bandwidths less than 300 Hz provide better harmonic rejection than higher bandwidths.

Operating Concepts Noise Reduction Techniques
Another difference between sweep-to-sweep averaging and variable IF bandwidth is the sweep time. Averaging displays the first complete trace faster but takes several sweeps to reach a fully averaged trace. IF bandwidth reduction lowers the noise floor in one sweep, but the sweep time may be slower. The difference in noise floor between a trace measured with a 3000 Hz IF bandwidth and with a 10 Hz IF bandwidth is illustrated by Figure 7-20.
Figure 7-20 IF Bandwidth Reduction

Another capability that can be used for effective noise reduction is the marker statistics function, which computes the average value of part or all of the formatted trace.
If your instrument is equipped with Option 085 (High Power System), another way of increasing dynamic range is to increase the input power to the test device using a booster amplifier.

Operating Concepts Measurement Calibration
Measurement calibration is an accuracy enhancement procedure that effectively removes the system errors that cause uncertainty in measuring a test device. It measures known standard devices, and uses the results of these measurements to characterize the system.
This section discusses the following topics:
A perfect measurement system would have infinite dynamic range, isolation, and directivity characteristics, no impedance mismatches in any part of the test setup, and flat frequency response. In any high frequency measurement there are measurement errors associated with the system that contribute uncertainty to the results. Parts of the measurement setup such as interconnecting cables and signal-separation devices (as well as the analyzer itself) all introduce variations in magnitude and phase that can mask the actual performance of the test device. Vector accuracy enhancement, also known as measurement calibration or error-correction, provides the means to simulate a nearly perfect measurement system.
For example, crosstalk due to the channel isolation characteristics of the analyzer can contribute an error equal to the transmission signal of a high-loss test device. For reflection measurements, the primary limitation of dynamic range is the directivity of the test setup. The measurement system cannot distinguish the true value of the signal reflected by the test device from the signal arriving at the receiver input due to leakage in the system. For both transmission and reflection measurements, impedance mismatches within the test setup cause measurement uncertainties that appear as ripples superimposed on the measured data.
Error-correction simulates an improved analyzer system. During the measurement calibration process, the analyzer measures the magnitude and phase responses of known standard devices, and compares the measurement with actual device data. The analyzer uses the results to characterize the system and effectively remove the system errors from the measurement data of a test device, using vector math capabilities internal to the network analyzer.
When you use a measurement calibration, the dynamic range and accuracy of the measurement are limited only by system noise and stability, connector repeatability, and the accuracy to which the characteristics of the calibration standards are known.

Operating Concepts Measurement Calibration
Network analysis measurement errors can be separated into systematic, random, and drift errors.
Correctable systematic errors are the repeatable errors that the system can measure. These are errors due to mismatch and leakage in the test setup, isolation between the reference and test signal paths, and system frequency response.
The system cannot measure and correct for the non-repeatable random and drift errors. These errors affect both reflection and transmission measurements. Random errors are measurement variations due to noise and connector repeatability. Drift errors include frequency drift, temperature drift, and other physical changes in the test setup between calibration and measurement.
The resulting measurement is the vector sum of the test device response plus all error terms. The precise effect of each error term depends upon its magnitude and phase relationship to the actual test device response.
In most high frequency measurements the systematic errors are the most significant source of measurement uncertainty. Since each of these errors can be characterized, their effects can be effectively removed to obtain a corrected value for the test device response. For the purpose of vector accuracy enhancement, these uncertainties are quantified as directivity, source match, load match, isolation (crosstalk), and frequency response (tracking). The description of each of these systematic errors follows.
Random and drift errors cannot be precisely quantified, so they must be treated as producing a cumulative uncertainty in the measured data.
Normally a device that can separate the reverse from the forward traveling waves (a directional bridge or coupler) is used to detect the signal reflected from the test device. Ideally the coupler would completely separate the incident and reflected signals, and only the reflected signal would appear at the coupled output, as shown in Figure 7-21a.
Directivity · Figure 7-21 Directivity

Operating Concepts Measurement Calibration
However, an actual coupler is not perfect, as shown in Figure 7-21b. A small amount of the incident signal appears at the coupled output due to leakage as well as reflection from the termination in the coupled arm. Also, reflections from the coupler output connector appear at the coupled output, adding uncertainty to the signal reflected from the device. The figure of merit for how well a coupler separates forward and reverse waves is directivity. The greater the directivity of the device, the better the signal separation. System directivity is the vector sum of all leakage signals appearing at the analyzer receiver input. The error contributed by directivity is independent of the characteristics of the test device and it usually produces the major ambiguity in measurements of low reflection devices.
Source match is defined as the vector sum of signals appearing at the analyzer receiver input due to the impedance mismatch at the test device looking back into the source, as well as to adapter and cable mismatches and losses. In a reflection measurement, the source match error signal is caused by some of the reflected signal from the test device being reflected from the source back toward the test device and re-reflected from the test device. This is illustrated in Figure 7-22. In a transmission measurement, the source match error signal is caused by reflection from the test device that is re-reflected from the source. Source match is most often given in terms of return loss in dB: thus the larger the number, the smaller the error.
Figure 7-22 Source Match
The error contributed by source match is dependent on the relationship between the actual input impedance of the test device and the equivalent match of the source. It is a factor in both transmission and reflection measurements. Source match is a particular problem in measurements where there is a large impedance mismatch at the measurement plane. (For example, reflection devices such as filters with stop bands.)
Source Match

Operating Concepts Measurement Calibration
Load match error results from an imperfect match at the output of the test device. It is caused by impedance mismatches between the test device output port and port 2 of the measurement system. Some of the transmitted signal is reflected from port 2 back to the test device as illustrated in Figure 7-23. A portion of this wave may be re-reflected to port 2, or part may be transmitted through the device in the reverse direction to appear at port 1. If the test device has low insertion loss (for example a filter pass band), the signal reflected from port 2 and re-reflected from the source causes a significant error because the test device does not attenuate the signal significantly on each reflection. Load match is usually given in terms of return loss in dB: thus the larger the number, the smaller the error.
Figure 7-23 Load Match
The error contributed by load match is dependent on the relationship between the actual output impedance of the test device and the effective match of the return port (port 2). It is a factor in all transmission measurements and in reflection measurements of two-port devices. The interaction between load match and source match is less significant when the test device insertion loss is greater than about 6 dB. However, source match and load match still interact with the input and output matches of the DUT, which contributes to transmission measurement errors. (These errors are largest for devices with highly reflective ports.)
Leakage of energy between analyzer signal paths contributes to error in a transmission measurement, much like directivity does in a reflection measurement. Isolation is the vector sum of signals appearing at the analyzer samplers due to crosstalk between the reference and test signal paths. This includes signal leakage within the test set and in both the RF and IF sections of the receiver.
The error contributed by isolation depends on the characteristics of the test device. Isolation is a factor in high-loss transmission measurements. However, analyzer system isolation is more than sufficient for most measurements, and correction for it may be unnecessary.
For measuring devices with high dynamic range, accuracy enhancement can provide improvements in isolation that are limited only by the noise floor. Generally, the isolation falls below the noise floor, therefore, when performing an isolation calibration you should use a noise reduction function such as averaging or reduce the IF bandwidth.
Load Match · Isolation (Crosstalk)

Operating Concepts Measurement Calibration
This is the vector sum of all test setup variations in which magnitude and phase change as a function of frequency. This includes variations contributed by signal-separation devices, test cables, adapters, and variations between the reference and test signal paths. This error is a factor in both transmission and reflection measurements.
For further explanation of systematic error terms and the way they are combined and represented graphically in error models, refer to the “Characterizing Microwave Systematic Errors” on page 7-41.
In a measurement of the reflection coefficient (magnitude and phase) of a test device, the measured data differs from the actual, no matter how carefully the measurement is made. Directivity, source match, and reflection signal path frequency response (tracking) are the major sources of error. See Figure 7-24.
Figure 7-24 Sources of Error in a Reflection Measurement
Frequency Response (Tracking) · One-Port Error Model

Operating Concepts Measurement Calibration
To characterize the errors, the reflection coefficient is measured by first separating the incident signal (I) from the reflected signal (R), then taking the ratio of the two values. See Figure 7-25. Ideally, (R) consists only of the signal reflected by the test device (S11A, for S11 actual).
Figure 7-25 Reflection Coefficient
However, all of the incident signal does not always reach the unknown. Refer to Figure 7-26. Some of (I) may appear at the measurement system input due to leakage through the test set or through a signal separation device. Also, some of (I) may be reflected by imperfect adapters between a signal separation device and the measurement plane. The vector sum of the leakage and the miscellaneous reflections is the effective directivity, EDF. Understandably, the measurement is distorted when the directivity signal combines vectorally with the actual reflected signal from the unknown, S11A.
Figure 7-26 Effective Directivity EDF
Since the measurement system test port is never exactly the characteristic impedance (50 ohms), some of the reflected signal bounces off the test port, or other impedance transitions further down the line, and back to the unknown, adding to the original incident signal (I). This effect causes the magnitude and phase of the incident signal to vary as a function of S11A and frequency. Leveling the source to produce a constant incident signal (I) reduces this error, but since the source cannot be exactly leveled at the test device input, leveling cannot eliminate all power variations.

Operating Concepts Measurement Calibration
This re-reflection effect and the resultant incident power variation are caused by the source match error, ESF as shown in Figure 7-27.
Figure 7-27 Source Match ESF
Frequency response (tracking) error is caused by variations in magnitude and phase flatness versus frequency between the test and reference signal paths. These are due mainly to coupler roll off, imperfectly matched samplers, and differences in length and loss between the incident and test signal paths. The vector sum of these variations is the reflection signal path tracking error, ERF as shown in Figure 7-28.
Figure 7-28 Reflection Tracking ERF
These three errors are mathematically related to the actual data, S11A, and measured data, S11M, by the following equation:
S 11 M E DF
1 ESF S11A – + =

Operating Concepts Measurement Calibration
If the value of these three "E" errors and the measured test device response were known for each frequency, this equation could be solved for S11A to obtain the actual test device response. Because each of these errors changes with frequency, their values must be known at each test frequency. These values are found by measuring the system at the measurement plane using three independent standards whose S11A is known at all frequencies.
The first standard applied is a "perfect load," which makes S11A = 0 and essentially measures directivity. See Figure 7-29. "Perfect load" implies a reflectionless termination at the measurement plane. All incident energy is absorbed. With S11A = 0 the equation can be solved for EDF, the directivity term. In practice, of course, the "perfect load" is difficult to achieve, although very good broadband loads are available in the compatible calibration kits.
Figure 7-29 "Perfect Load" Termination
Since the measured value for directivity is the vector sum of the actual directivity plus the actual reflection coefficient of the "perfect load," any reflection from the termination represents an error. System effective directivity becomes the actual reflection coefficient of the near "perfect load" as shown in Figure 7-30. In general, any termination having a return loss value greater than the uncorrected system directivity reduces reflection measurement uncertainty.

Operating Concepts Measurement Calibration
Figure 7-30 Measured Effective Directivity
Next, a short circuit termination whose response is known to a very high degree is used to establish another condition as shown in Figure 7-31.
Figure 7-31 Short Circuit Termination
The open circuit gives the third independent condition. In order to accurately model the phase variation with frequency due to fringing capacitance from the open connector, a specially designed shielded open circuit is used for this step. (The open circuit capacitance is different with each connector type.) Now the values for EDF, directivity, ESF, source match, and ERF, reflection frequency response, are computed and stored. See Figure 7-32.

Operating Concepts Measurement Calibration
Figure 7-32 Open Circuit Termination
This completes the calibration procedure for one port devices.
Now the unknown is measured to obtain a value for the measured response, S11M, at each frequency. Refer to Figure 7-33.
Figure 7-33 Measured S11
This is the one-port error model equation solved for S11A. Since the three errors and S11M are now known for each test frequency, S11A can be computed as follows:
S 11 A
ESF S11M EDF – ( ) ERF + =
For reflection measurements on two-port devices, the same technique can be applied, but the test device output port must be terminated in the system characteristic impedance. This termination should have as low a reflection coefficient as the load used to determine directivity. The additional reflection error caused by an improper termination at the test device’s output port is not incorporated into the one-port error model.
Device Measurement

Operating Concepts Measurement Calibration
Two-Port Error Model (ES Models Only)
The error model for measurement of the transmission coefficients (magnitude and phase) of a two-port device is derived in a similar manner. The potential sources of error are frequency response (tracking), source match, load match, and isolation as shown in Figure 7-34. These errors are effectively removed using the full two-port error model.
Figure 7-34 Major Sources of Error
The transmission coefficient is measured by taking the ratio of the incident signal (I) and the transmitted signal (T). Refer to Figure 7-35. Ideally, (I) consists only of power delivered by the source, and (T) consists only of power emerging at the test device output.
Figure 7-35 Transmission Coefficient

Operating Concepts Measurement Calibration
As in the reflection model, source match can cause the incident signal to vary as a function of test device S11A. Also, since the test setup transmission return port is never exactly the characteristic impedance, some of the transmitted signal is reflected from the test set port 2, and from other mismatches between the test device output and the receiver input, to return to the test device. A portion of this signal may be re-reflected at port 2, thus affecting S21M, or part may be transmitted through the device in the reverse direction to appear at port 1, thus affecting S11M. This error term, which causes the magnitude and phase of the transmitted signal to vary as a function of S22A, is called load match, ELF. See Figure 7-36.
Figure 7-36 Load Match ELF
The measured value, S21M, consists of signal components that vary as a function of the relationship between ESF and S11A as well as ELF and S22A, so the input and output reflection coefficients of the test device must be measured and stored for use in the S21A error-correction computation. Thus, the test setup is calibrated as described for reflection to establish the directivity, EDF, source match, ESF, and reflection frequency response, ERF, terms for reflection measurements on both ports.
Now that a calibrated port is available for reflection measurements, the thru is connected and load match, ELF, is determined by measuring the reflection coefficient of the thru connection.
Transmission signal path frequency response is then measured with the thru connected. The data is corrected for source and load match effects, then stored as transmission frequency response, ETF.

Operating Concepts Measurement Calibration
NOTE It is very important that the exact electrical length of the thru be known. Most calibration kits assume a zero length thru. For some connection types such as Type-N, this implies one male and one female port. If the test system requires a non-zero length thru, for example, one with two male test ports, the exact electrical delay of the thru adapter must be used to modify the built-in calibration kit definition of the thru.
Isolation, EXF, represents the part of the incident signal that appears at the receiver without actually passing through the test device. See Figure 7-37. Isolation is measured with the test set in the transmission configuration and with terminations installed at the points where the test device will be connected. Since isolation can be lower than the noise floor, it is best to increase averaging by at least a factor of four during the isolation portion of the calibration. The softkey under the menu allows a calibration sequence to resume after a change to the averaging factor.
RESUME CAL SEQUENCE Cal
If the leakage falls below the noise floor, it is best to increase averaging before calibration. In this case, omitting isolation is better than measuring the isolation standards without increasing the averaging factor.
Figure 7-37 Isolation EXF
Thus there are two sets of error terms, forward and reverse, with each set consisting of six error terms, as follows:

Operating Concepts Measurement Calibration
The analyzer’s test set can measure both the forward and reverse characteristics of the test device without you having to manually remove and physically reverse the device. A full two-port error model illustrated in Figure 7-38. This illustration depicts how the analyzer effectively removes both the forward and reverse error terms for transmission and reflection measurements.
Figure 7-38 Full Two-Port Error Model
A full two-port error model equations for all four S-parameters of a two-port device is shown in Figure 7-39. Note that the mathematics for this comprehensive model use all forward and reverse error terms and measured values. Thus, to perform full error-correction for any one parameter, all four S-parameters must be measured.
Applications of these error models are provided in the calibration procedures described in Chapter 5 , “Optimizing Measurement Results.”
Enhanced Response Calibration Error Model Enhanced response calibration uses the same error model as the forward configuration portion of Figure 7-38. In the response portion, the source and load match effects are fully-accounted for giving the same accuracy to the forward tracking term (ETF) as the two-port calibration. During the measurement, the enhanced response calibration performs a correction which is mathematically the same as setting the values of ELF, ELR, ESR, EXR, EDR to zero (0) and the values of ERR and ETR to one (1) in the equations for S11 and S21 shown in Figure 7-39.

Operating Concepts Measurement Calibration
Figure 7-39 Full Two-Port Error Model Equations
In addition to the errors removed by accuracy enhancement, other systematic errors exist due to limitations of dynamic accuracy, test set switch repeatability, and test cable stability. These, combined with random errors, also contribute to total system measurement uncertainty. Therefore, after accuracy enhancement procedures are performed, residual measurement uncertainties remain.
The uncorrected performance of the analyzer is sufficient for many measurements. However, the next three illustrations show the improvements that can be made in measurement accuracy by using a more complete calibration routine. See Figure 7-40, Figure 7-41, and Figure 7-42.

Operating Concepts Measurement Calibration
Figure 7-40a shows a measurement in log magnitude format with a response calibration only. Figure 7-40b shows the improvement in the same measurement using an S11 one-port calibration. Figure 7-41a shows the measurement on a Smith chart with response calibration only, and Figure 7-41b shows the same measurement with an S11 one-port calibration.
Figure 7-40 Response versus S11 1-Port Calibration on Log Magnitude Format

Figure 7-41 Response versus S11 1-Port Calibration on Smith Chart


Operating Concepts Measurement Calibration
The response of a device in a log magnitude format is shown in Figure 7-42. Figure 7-42a shows the response using a response calibration and Figure 7-42b the response using a full two-port calibration.
Figure 7-42 Response versus Full Two-Port Calibration

Operating Concepts Calibration Routines
There are twelve different error terms for a two-port measurement that can be corrected by accuracy enhancement in the analyzer. These are directivity, source match, load match, isolation, reflection tracking, and transmission tracking, each in both the forward and reverse direction. The analyzer has several different measurement calibration routines to characterize one or more of the systematic error terms and remove their effects from the measured data.
The calibrate menu allows you to perform the measurement calibration routines. These procedures range from a simple frequency response calibration to a full two-port calibration that effectively removes all twelve error terms.
The response calibration, activated by pressing the softkey within the calibrate menu, provides a normalization of the test setup for reflection or transmission measurements. This calibration procedure may be adequate for measurement of well matched devices. This is the simplest error-correction to perform, and should be used when extreme measurement accuracy is not required.
Response and Isolation Calibration
The response and isolation calibration, activated by pressing the softkey within the calibrate menu, provides a normalization for frequency response and crosstalk errors in transmission measurements, or frequency response and directivity errors in reflection measurements. This procedure may be adequate for measurement of well matched high-loss devices.
Enhanced Response Calibration
The enhanced response calibration, activated by pressing the softkey within the calibrate menu, provides a one-port calibration to correct for directivity, source match, and frequency response for reflection measurements, and corrects for source match as well as frequency response for transmission measurements.
For ES models, enhanced response calibration improves accuracy in transmission measurements compared to a response calibration or a response and isolation calibration, but it is not as accurate as a full 2-port calibration. For ET models, this calibration provides the best accuracy for measurements with a T/R test set.

Operating Concepts Calibration Routines
The enhanced reflection calibration is activated by selecting under the menu.
ENH. REFL. ON off ENHANCED RESPONSE
The enhanced reflection calibration effectively removes load match error from the enhanced response calibration performed on a bilateral device. A bilateral device has an identical forward (S21) and reverse transmission (S12) response. Most passive devices (such as filters, attenuators, or switches) are bilateral. Some passive devices (circulators, isolators) and most active devices do not have identical forward and reverse transmission responses and enhanced reflection calibration will not work with these devices.
The S11 and S22 one-port calibration procedures, activated by pressing the or
softkey within the calibrate menu, provide directivity, source match, and frequency response vector error-correction for reflection measurements. These procedures provide high accuracy reflection measurements of one-port devices or properly terminated two-port devices.
Full Two-Port Calibration (ES Models Only)
The full two-port calibration, activated by pressing the softkey within the calibrate menu, provides directivity, source match, load match, isolation, and frequency response vector error-correction, in both forward and reverse directions, for transmission and reflection measurements of two-port devices. This calibration provides the best magnitude and phase measurement accuracy for both transmission and reflection measurements of two-port devices, and requires an S-parameter test set.
In this type of calibration, both forward and reverse measurements must be made. You have the option of setting the ratio of the number of forward (or reverse) sweeps versus the number of reverse (or forward) sweeps. To access this function, press and enter the number of sweeps desired.
Cal MORE TESTSET SW
The TRL*/LRM* two-port calibration, activated by pressing the softkey within the calibration menu, provides the ability to make calibrations using the TRL or LRM method. True TRL/LRM calibration is available on analyzers equipped with Option 400, the Four Sampler Test Set. For more information, refer to “TRL*/LRM* Calibration (ES Models Only)” on page 7-67.
Enhanced Reflection Calibration

Operating Concepts Calibration Routines
The E-Cal calibration menu is activated by pressing in the calibration menu.
The E-Cal (Electronic Calibration) system determines systemic errors of the analyzer through a one-time connection of an E-Cal module to the network analyzer ports. The random error of connector repeatability is reduced substantially through a one-time connection when compared to frequent connections and disconnections of the conventional short/open/load methods.

Operating Concepts Modifying Calibration Kits
Modifying calibration kits is necessary only if unusual standards (such as in TRL*) are used or the very highest accuracy is required. Unless a calibration kit model is provided with the calibration devices used, a solid understanding of error-correction and the system error model are absolutely essential to making modifications. You may use modifications to a predefined calibration kit by modifying the kit and saving it as a user kit. The original predefined calibration kit will remain unchanged.
Before attempting to modify calibration standard definitions, you should read Application Note 8510-5A to improve your understanding of modifying calibration kits. The part number of this application note is 5956-4352. Although the application note is written for the 8510 family of network analyzers, it also applies to this network analyzer.
Several situations exist that may require a user-defined calibration kit:
The following are definitions of terms:

Operating Concepts Modifying Calibration Kits
The following steps are used to modify or define a user kit:
The softkey in the cal kit menu accesses the modify calibration kit menu. This leads in turn to additional series of menus associated with modifying calibration kits. The following is a description of the softkeys located within this menu:

Operating Concepts Modifying Calibration Kits
softkey, if it is to be used later.
Standard definition is the process of mathematically modeling the electrical characteristics (delay, attenuation, and impedance) of each calibration standard. These electrical characteristics (coefficients) can be mathematically derived from the physical dimensions and material of each calibration standard, or from its actual measured response. The parameters of the standards can be listed in Table 7-1.
Table 7-1 Standard Definitions
Calibration Kit Label
C0c ×10−15 F
C1c ×10−27 F/Hz
C2c ×10−36 F/Hz2
C3c ×10−45 F/Hz3
FIXEDd SLIDING or OFFSET
TERMe IMPED Ω
OFFSET FREQ (GHz) COAX or WG
NO. TYPE DELAY s
MIN MAX
a. Ensure system Z0 of network analyzer is set to this value.
b. Open, short, load, delay/thru, or arbitrary impedance.
c. Open standard types only.
d. Load or arbitrary impedance only.
e. Arbitrary impedance only, device terminating impedance.
Each standard must be identified as one of five "types": open, short, load, delay/thru, or arbitrary impedance.

Operating Concepts Modifying Calibration Kits
After a standard number is entered, selection of the standard type will present one of five menus for entering the electrical characteristics (model coefficients) corresponding to that standard type, such as . These menus are tailored to the current type, so that only characteristics applicable to the standard type can be modified.
The following is a description of the softkeys located within the define standard menu:
As a reflection standard, an open termination offers the advantage of broadband frequency coverage. At RF and microwave frequencies, however, an open rarely has perfect reflection characteristics because fringing capacitance effects cause phase shift that varies with frequency. This can be observed in measuring an open termination after calibration, when an arc in the lower right circumference of the Smith chart indicates capacitive reactance. These effects are impossible to eliminate, but the calibration kit models include the open termination capacitance at all frequencies for compatible calibration kits. The capacitance model is a cubic polynomial, as a function of frequency, where the polynomial coefficients are user-definable. The capacitance model equation is:
C = (C0) + (C1 × F) + (C2 × F2) + (C3 × F3)
where F is the measurement frequency.
The terms in the equation are defined with the specify open menu as follows:
— allows you to enter the C0 term, which is the constant term of the cubic polynomial and is scaled by 10−15 Farads.
— allows you to enter the C1 term, expressed in F/Hz (Farads/Hz) and scaled by 10−27.
— allows you to enter the C2 term, expressed in F/Hz2 and scaled by 10−36.
— allows you to enter the C3 term, expressed in F/Hz3 and scaled by 10−45.
— defines the load as a fixed (not sliding) load.
— defines the load as a sliding load. When such a load is measured during calibration, the analyzer will prompt for several load positions, and calculate the ideal load value from it.
— defines the load as being offset.

Operating Concepts Modifying Calibration Kits
— allows you to specify the (arbitrary) impedance of the standard, in ohms.
— defines the load as a fixed (not sliding) load.
— defines the load as a sliding load. When such a load is measured during calibration, the analyzer will prompt for several load positions, and calculate the ideal load value from it. Normally, arbitrary impedance standards are fixed rather than sliding.
Any standard type can be further defined with offsets in delay, loss, and standard impedance; assigned minimum or maximum frequencies over which the standard applies, and defined as coax or waveguide. The softkey accesses the specify offset menu (described next).
The softkey allows you to define a distinct label for each standard, so that the analyzer can prompt you with explicit standard labels during calibration (such as SHORT). The function is similar to defining a display title, except that the label is limited to ten characters.
After each standard is defined, including offsets, the softkey will terminate the standard definition.
The specify offset menu allows additional specifications for a user-defined standard. Features specified in this menu are common to all five types of standards.
Offsets may be specified with any standard type. This means defining a uniform length of transmission line to exist between the standard being defined and the actual measurement plane. (Example: a waveguide short circuit terminator, offset by a short length of waveguide.) For reflection standards, the offset is assumed to be between the measurement plane and the terminating element of the standard (one-way only). For transmission standards, the offset is assumed to exist between the two reference planes (in effect, the offset is the thru). For both reflection and transmission, the offset is entered as a one-way offset. Three characteristics of the offset can be defined: its delay (length), loss, and impedance.
In addition, the frequency range over which a particular standard is valid can be defined with a minimum and maximum frequency. This is particularly important for a waveguide standard, since the minimum frequency is used to define the waveguide cutoff frequency. Note that several band-limited standards can together be defined as the same "class" (see “Specify Class Menu” on page 7-62). Then, if a measurement calibration is performed over a frequency range exceeding a single standard, additional standards can be used for each portion of the range.
Lastly, the standard must be defined as either coaxial or waveguide. If it is waveguide, dispersion effects are calculated automatically and included in the standard model.
Specify Offset Menu

Operating Concepts Modifying Calibration Kits
The following is a description of the softkeys located within the specify offset menu:
In coax, group delay is considered constant. In waveguide, however, group delay is dispersive, that is, it changes significantly as a function of frequency. Hence, for a waveguide standard, offset delay must be defined as though it were a TEM wave (without dispersion).
OFFSET DELAY
OFFSET DELAY
This menu allows you to label (reference) individual standards during the menu-driven measurement calibration sequence. The labels are user-definable using a character set shown on the display that includes letters, numbers, and some symbols, and they may be up to ten characters long. The analyzer will prompt you to connect standards using these labels, so they should be meaningful to you, and distinct for each standard.
By convention, when sexed connector standards are labeled male (m) or female (f), the designation refers to the test port connector sex, not the connector sex of the standard.
Once a standard has been defined, it must be assigned to a standard "class." This is a group of from one to seven standards that is required to calibrate for a single error term. The standards within a single class can be assigned to the locations listed in Table 7-2 according to their standard reference numbers.
Label Standard Menu · Specify Class Menu

Operating Concepts Modifying Calibration Kits
A class often consists of a single standard, but may be composed of more than one standard if band-limited standards are used. For example, if there were two load standards—a fixed load for low frequencies, and a sliding load for high frequencies—then that class would have two standards.
Table 7-2 Standard Class Assignments
Class Standard Reference Numbers Standard Class Label 1 2 3 4 5 6 7 8
S 11 B
S 11 C
S 22 A
S 22 B
S 22 C
TRL line or match
The number of standard classes required depends on the type of calibration being performed, and is identical to the number of error terms corrected. A response calibration requires only one class, and the standards for that class may include an open, or short, or thru. A 1-port calibration requires three classes. A full 2-port calibration requires 10 classes, not including two for isolation.
The number of standards that can be assigned to a given class may vary from none (class not used) to one (simplest class) to seven. When a certain class of standards is required during calibration, the analyzer will display the labels for all the standards in that class (except when the class consists of a single standard). This does not, however, mean that all standards in a class must be measured during calibration. Unless band-limited standards are used, only a single standard per class is required.

Operating Concepts Modifying Calibration Kits
Each class can be given a user-definable label as described under label class menus.
Standards are assigned to a class simply by entering the standard’s reference number (established while defining a standard) under a particular class. The following is a description of the softkeys located within the specify class menu:

Operating Concepts Modifying Calibration Kits
The label class menus are used to define meaningful labels for the calibration classes. These then become softkey labels during a measurement calibration. Labels can be up to ten characters long.
This softkey within the modify cal kit menu, accesses this menu. It is identical to the label class menu and the label standard menu previously described. It allows definition of a label up to eight characters long.
After a new calibration kit has been defined, be sure to specify a label for it. Choose a label that describes the connector type of the calibration devices. This label will then appear in the softkey label in the correction menu and the label in the select cal kit menu. It will be saved with calibration sets.
Verify Performance
Once a measurement calibration has been generated with a user-defined calibration kit, its performance should be checked before making device measurements. To check the accuracy that can be obtained using the new calibration kit, a device with a well-defined frequency response (preferably unlike any of the standards used) should be measured. The verification device must not be one of the calibration standards: measurement of one of these standards is merely a measure of repeatability.
To achieve more complete verification of a particular measurement calibration, accurately known verification standards with a diverse magnitude and phase response should be used. National standard traceable or Agilent standards are recommended to achieve verifiable measurement accuracy.
The calibration kit, along with any calibration data and other instrument state information, can be saved to an ISTATE file on a floppy disk. To save a modified calibration kit with an instrument state, press:
Cal CAL KIT SELECT CAL KIT USER KIT
Label Class Menu · Label Kit Menu · Save/Recall SAVE STATE

Operating Concepts Modifying Calibration Kits
To modify a calibration kit from the calibration kit selection menu, press:
Cal CAL KIT SELECT CAL KIT MODIFY KIT DONE (MODIFIED)
To save the modified calibration kit, press:
Cal CAL KIT SELECT CAL KIT USER KIT SAVE USER KIT Save/Recall SAVE STATE
Ensure that is underlined before saving the modified user kit.
or

Operating Concepts TRL*/LRM* Calibration (ES Models Only)
The network analyzer has the capability of making calibrations using the "TRL" (thru-reflect-line) method. This section contains information on the following subjects:
— Requirements for TRL Standards
This method is convenient in that calibration standards can be fabricated for a specific measurement environment, such as a transistor test fixture or microstrip. Microstrip devices in the form of chips, MMIC's, packaged transistors, or beam-lead diodes cannot be connected directly to the coaxial ports of the analyzer. The device under test (DUT) must be physically connected to the network analyzer by some kind of transition network or fixture. Calibration for a fixtured measurement in microstrip presents additional difficulties.
A calibration at the coaxial ports of the network analyzer removes the effects of the network analyzer and any cables or adapters before the fixture; however, the effects of the fixture itself are not accounted for. An in-fixture calibration is preferable, but high-quality short-open-load- thru (SOLT) standards may not be readily available to allow a conventional full 2-port calibration of the system at the desired measurement plane of the device. In microstrip, a short circuit is inductive, an open circuit radiates energy, and a high-quality purely resistive load is difficult to produce over a broad frequency range. The Thru-Reflect-Line (TRL) 2-port calibration is an alternative to the traditional SOLT Full 2-port calibration technique that utilizes simpler, more convenient standards for device measurements in the microstrip environment.
For coaxial, waveguide and other environments where high-quality impedance standards are readily available, the traditional short, open, load, thru (SOLT) method provides the most accurate results since all of the significant systematic errors are reduced. This method is implemented in the form of the S11 1-port, S22 1-port, and full 2-port calibration selections.
In all measurement environments, you must provide calibration standards for the desired calibration to be performed. The advantage of TRL is that only three standards need to be characterized as opposed to 4 in the traditional open, short, load, and thru full 2-port calibrations. Further, the requirements for characterizing the T, R, and L standards are less stringent and these standards are more easily fabricated.
— TRL Options

Operating Concepts TRL*/LRM* Calibration (ES Models Only)
TRL* calibration is a modified form of TRL calibration. It is adapted for a receiver with three samplers instead of four samplers. The TRL* calibration is not as accurate as the TRL calibration because it cannot isolate the source match from the load match, so it assumes load match and source match are equal. However, option 400 on the ES model adds the fourth sampler and provides for a true TRL calibration.
The TRL*/LRM* calibration used in the analyzer relies on the characteristic impedance of simple transmission lines rather than on a set of discrete impedance standards. Since transmission lines are relatively easy to fabricate (in a microstrip, for example), the impedance of these lines can be determined from the physical dimensions and substrate’s dielectric constant.
Figure 7-43 Functional Block Diagram for a 2-Port Error-Corrected

For the analyzer TRL* 2-port calibration, a total of 10 measurements are made to quantify eight unknowns (not including the two isolation error terms). Assume the two transmission leakage terms, EXF and EXR, are measured using the conventional technique. The eight TRL error terms are represented by the error adapters shown in Figure 7-43. Although this error model is slightly different from the traditional Full 2-port 12-term model, the conventional error terms may be derived from it. For example, the forward reflection tracking (ERF) is represented by the product of ε10 and ε01.
TRL* Error Model · Measurement System

Operating Concepts TRL*/LRM* Calibration (ES Models Only)
Also notice that the forward source match (ESF) and reverse load match (ELR) are both represented by ε11, while the reverse source match (ESR) and forward load match (ELF) are both represented by ε22. In order to solve for these eight unknown TRL error terms, eight linearly independent equations are required.
The first step in the TRL* 2-port calibration process is the same as the transmission step for a Full 2-port calibration. For the thru step, the test ports are connected together directly (zero length thru) or with a short length of transmission line (non- zero length thru) and the transmission frequency response and port match are measured in both directions by measuring all four S-parameters.
For the reflect step, identical high reflection coefficient standards (typically open or short circuits) are connected to each test port and measured (S11 and S22).
For the line step, a short length of transmission line (different in length from the thru) is inserted between port 1 and port 2 and again the frequency response and port match are measured in both directions by measuring all four S-parameters.
In total, ten measurements are made, resulting in ten independent equations. However, the TRL error model has only eight error terms to solve for. The characteristic impedance of the line standard becomes the measurement reference and, therefore, has to be assumed ideal (or known and defined precisely).
At this point, the forward and reverse directivity (EDF and EDR), transmission tracking (ETF and ETR), and reflection tracking (ERF and ERR) terms may be derived from the TRL error terms. This leaves the isolation (EXF and EXR), source match (ESF and ESR) and load match (ELF and ELR) terms to discuss.
Two additional measurements are required to solve for the isolation terms (EXF and EXR). Isolation is characterized in the same manner as the Full 2-port calibration. Forward and reverse isolation are measured as the leakage (or crosstalk) from port 1 to port 2 with each port terminated. The isolation part of the calibration is generally only necessary when measuring high loss devices (greater than 70 dB).
Isolation

Operating Concepts TRL*/LRM* Calibration (ES Models Only)
Figure 7-44 8-term TRL (or TRL*) Error Model and Generalized Coefficients

Source match and load match
A TRL calibration assumes a perfectly balanced test set architecture as shown by the term which represents both the forward source match (ESF) and reverse load match (ELR), and by the ε22 term which represents both the reverse source match (ESR) and forward load match (ELF). However, in any switching test set, the source and load match terms are not equal because the transfer switch presents a different terminating impedance as it is changed between port 1 and port 2.
Because the standard network analyzer is based on a three-sampler receiver architecture, it is not possible to differentiate the source match from the load match terms. The terminating impedance of the switch is assumed to be the same in either direction. Therefore, the test port mismatch cannot be fully corrected. An assumption is made that:
forward source match (ESF ) = reverse load match (ELR) = ε11
reverse source match (ESR ) = forward load match (ELF) = ε22
For a fixture, TRL* can eliminate the effects of the fixture’s loss and length, but does not completely remove the effects due to the mismatch of the fixture.

Operating Concepts TRL*/LRM* Calibration (ES Models Only)
A technique that can be used to improve the raw test port mismatch is to add high quality fixed attenuators. The effective match of the system is improved because the fixed attenuators usually have a return loss that is better than that of the network analyzer. Additionally, the attenuators provide some isolation of reflected signals. The attenuators also help to minimize the difference between the port source match and load match, making the error terms more equivalent.
With the attenuators in place, the effective port match of the system is improved so that the mismatch of the fixture transition itself dominates the measurement errors after a calibration.
Figure 7-45 Typical Measurement Set up
If the device measurement requires bias, it will be necessary to add external bias tees between the fixed attenuators and the fixture. The internal bias tees of the analyzer will not pass the bias properly through the external fixed attenuators. Be sure to calibrate with the external bias tees in place (no bias applied during calibration) to remove their effect from the measurement.
Because the bias tees must be placed after the attenuators, they essentially become part of the fixture. Their mismatch effects are the same for source match and load match, so the TRL CAL routine will correct for their effects. Although the fixed attenuators improve the raw mismatch of the network analyzer system, they also degrade the overall measurement dynamic range.
This effective mismatch of the system after calibration has the biggest effect on reflection measurements of highly reflective devices. Likewise, for well matched devices, the effects of mismatch are negligible. This can be shown by the following approximation:
Reflection magnitude uncertainty = ED + ERS11 + ES(S11)2 + ELS21S12

Operating Concepts TRL*/LRM* Calibration (ES Models Only)
Transmission magnitude uncertainty = EX + ETS21 + ESS11S21 + ELS22S21
where:
ED = effective directivity
ER = effective reflection tracking
ES = effective source match
EL = effective load match
EX = effective crosstalk
ET = effective transmission tracking
Sxx= S-parameters of the device under test
The TRL implementation with Option 400 requires a total of fourteen measurements to quantify ten unknowns as opposed to only a total of twelve measurements for TRL*. (Both include the two isolation error terms.)
Because of the four-sampler receiver architecture of Option 400, additional correction of the source match and load match terms is achieved by measuring the ratio of the two “reference” receivers during the thru and line steps. These measurements characterize the impedance of the switch and associated hardware in both the forward and reverse measurement configurations. They are then used to modify the corresponding source and load match terms (for both forward and reverse).
The Option 400 configuration with TRL established a higher performance calibration method over TRL* when making in-fixture measurements, because all significant error terms are systematically reduced. With TRL*, the source and load match terms are essentially that of the raw, “uncorrected” performance of the hardware.
Requirements for TRL Standards
When building a set of TRL standards for a microstrip or fixture environment, the requirements for each of these standard types must be satisfied.
THRU (Zero length) • No loss. Characteristic impedance (Z0 ) need not be known.
THRU (Non-zero length) • Z0 of the thru must be the same as the line. (If they are not the same, the average impedance is used.)
Types Requirements

Operating Concepts TRL*/LRM* Calibration (ES Models Only)
REFLECT • Reflection coefficient Γ magnitude is optimally 1.0, but need not be known.
LINE/MATCH (LINE) • Z0 of the line establishes the reference impedance of the measurement (i.e. S11= S22 = 0). The calibration impedance is defined to be the same as Z0 of the line. If the Z0 is known but not the desired value (i.e., not equal to 50 Ω), the SYSTEMS Z0 selection under the TRL/LRM options menu is used.
LINE/MATCH (MATCH) • Z0 of the match establishes the reference impedance of the measurement.

Operating Concepts TRL*/LRM* Calibration (ES Models Only)
Fabricating and defining calibration standards for TRL/LRM
When calibrating a network analyzer, the actual calibration standards must have known physical characteristics. For the reflect standard, these characteristics include the offset in electrical delay (seconds) and the loss (ohms/second of delay). The characteristic impedance, , is not used in the calculations in that it is determined by the line standard. The reflection coefficient magnitude should optimally be 1.0, but need not be known since the same reflection coefficient magnitude must be applied to both ports.
The thru standard may be a zero-length or known length of transmission line. The value of length must be converted to electrical delay, just like that done for the reflect standard. The loss term must also be specified.
The line standard must meet specific frequency related criteria, in conjunction with the length used by the thru standard. In particular, the insertion phase of the line must not be the same as the thru. The optimal line length is 1/4 wavelength (90 degrees) relative to a zero length thru at the center frequency of interest, and between 20 and 160 degrees of phase difference over the frequency range of interest. (Note: these phase values can be ±N × 180 degrees where N is an integer.) If two lines are used (LRL), the difference in electrical length of the two lines should meet these optimal conditions. Measurement uncertainty will increase significantly when the insertion phase nears zero or is an integer multiple of 180 degrees, and this condition is not recommended.
For a transmission media that exhibits linear phase over the frequency range of interest, the following expression can be used to determine a suitable line length of one-quarter wavelength at the center frequency (which equals the sum of the start frequency and stop frequency divided by 2):
Electrical length cm ( ) LINE 0 length THRU – ( ) =
Electrical length cm ( )
f1 MHz ( ) f2 MHz ( ) + =
let:
f1 = 1000 MHz
f2 = 2000 MHz
VF = Velocity Factor = 1 (for this example)
Thus, the length to initially check is 5 cm.
Next, use the following to verify the insertion phase at f1 and f2:
Phase degrees ( ) 360 f × l × ( ) v =

Operating Concepts TRL*/LRM* Calibration (ES Models Only)
where:
l = length of line
v = velocity = speed of light × velocity factor
which can be reduced to the following using frequencies in MHz and length in centimeters:
Phase degrees ( ) approx 0.012 f MHz ( ) × l cm ( ) × VF =
So for an air line (velocity factor approximately 1) at 1000 MHz, the insertion phase is 60 degrees for a 5 cm line; it is 120 degrees at 2000 MHz. This line would be a suitable line standard.
For microstrip and other fabricated standards, the velocity factor is significant. In those cases, the phase calculation must be divided by that factor. For example, if the dielectric constant for a substrate is 10, and the corresponding "effective" dielectric constant for microstrip is 6.5, then the "effective" velocity factor equals 0.39 (1 ÷ square root of 6.5).
Using the first equation with a velocity factor of 0.39, the initial length to test would be 1.95 cm. This length provides an insertion phase at 1000 MHz of 60 degrees; at 2000 MHz, 120 degrees (the insertion phase should be the same as the air line because the velocity factor was accounted for when using the first equation).
Another reason for showing this example is to point out the potential problem in calibrating at low frequencies using TRL. For example, one-quarter wavelength is
Length cm ( ) 7500 VF × fc =
where:
Thus, at 50 MHz,
Length cm ( ) 7500 50 MHz ( ) 150 cm or 1.5 m = =
Such a line standard would not only be difficult to fabricate, but its long term stability and usability would be questionable as well.
Thus at lower frequencies or very broad band measurements, fabrication of a "match" or termination may be deemed more practical. Since a termination is, in essence, an infinitely long transmission line, it fits the TRL model mathematically, and is sometimes referred to as a "TRM" calibration.
f = frequency

Operating Concepts TRL*/LRM* Calibration (ES Models Only)
The TRM calibration technique is related to TRL with the difference being that it bases the characteristic impedance of the measurement on a matched ZO termination instead of a transmission line for the third measurement standard. Like the TRL thru standard, the TRM THRU standard can either be of zero length or non-zero length. The same rules for thru and reflect standards used for TRL apply for TRM.
TRM has no inherent frequency coverage limitations which makes it more convenient in some measurement situations. Additionally, because TRL requires a different physical length for the thru and the line standards, its use becomes impractical for fixtures with contacts that are at a fixed physical distance from each other.
For information on how to modify calibration constants for TRL*/LRM*, and how to perform a TRL or TRM calibration, refer to “Calibrating for Non-Coaxial Devices (ES Analyzers Only)” on page 6-50.
The softkey accesses the TRL/LRM options menu. There are two selections under this menu:
The characteristic impedance used during the calibration can be referenced to either the line (or match) standard ( ) or to the system ( ). The analyzer defaults to a calibration impedance that is equal to the line (or match) standard.
When the is selected, the impedance of the line (or match) standard is assumed to match the system impedance exactly (the line standard is reflectionless). After a calibration, all measurements are referenced to the impedance of the line standard. For example, when the line standard is remeasured, the response will appear at the center of the Smith chart. When is selected, the values entered for (under menu) and (within the define standard menu) are ignored.
CAL ZO: LINE ZO SET ZO
is selected when the desired measurement impedance differs from the impedance of the line standard. This requires a knowledge of the exact value of the Z0
of the line. The system reference impedance is set using under the calibration menu. The actual impedance of the line is set by entering the real part of the line impedance as the within the define standard menu. For example, if the line was known to have a characteristic impedance of 51 Ω ( = 51 Ω), it could still be used to calibrate for a 50 Ω measurement ( = 50 Ω). After a calibration, all measurements would be referenced to 50 Ω, instead of 51 Ω. When the line standard is remeasured, the center of the Smith chart is at the current value of (in this case, 50 Ω). Since only one value of offset Z0 can be selected for the line standard, the value of Z0 should be a constant value over the frequency range of interest in order to be meaningful.
OFFSET ZO
OFFSET ZO SET ZO
SET ZO
TRL Options · Cal OFFSET ZO

Operating Concepts TRL*/LRM* Calibration (ES Models Only)
The location of the reference plane is determined by the selection of and . By default, the reference plane is set with the thru standard which must have a known insertion phase or electrical length. If a non-zero length thru is specified to have zero delay, the reference plane will be established in the middle of the thru. The reflect standard may be used to set the reference plane instead of the thru provided the phase response (offset delay, reactance values and standard type) of the reflect standard is known and is specified in the calibration kit definition.
Dispersion occurs when a transmission medium exhibits a variable propagation or phase velocity as a function of frequency. The result of dispersion is a non-linear phase shift versus frequency, which leads to a group delay which is not constant. Fortunately, the TRL calibration technique accounts for dispersive effects of the test fixture up to the calibration plane, provided that:
When a non-zero length thru is used to set the reference plane, it should be defined as having zero length in the TRL standards definition, even though it has physical length. The actual electrical length of the thru standard must then be subtracted from the actual electrical length of each line standard in the TRL calibration kit definition. The device must then be mounted between two short lengths of transmission line so that each length is exactly one-half of the length of the non-zero length thru standard. In this configuration, the measurement will be properly calibrated up to the point of the device.

Operating Concepts GPIB Operation
This section contains information on the following topics:
This key is allows you to return the analyzer to local (front panel) operation from remote (computer controlled) operation. This key will also abort a test sequence or hardcopy print/plot. In this local mode, with a controller still connected on GPIB, you can operate the analyzer manually (locally) from the front panel. This is the only front panel key that is not disabled when the analyzer is remotely controlled over GPIB by a computer. The exception to this is when local lockout is in effect: this is a remote command that disables the key, making it difficult to interfere with the analyzer while it is under computer control.
In addition, the key accesses the GPIB menu, where you can set the controller mode, and to the address menu, where you can enter the GPIB addresses of peripheral devices and select plotter/printer ports. You can also set the mode of the parallel port here.
The GPIB menu consists of the following softkeys:
TALKER/LISTENER
USE PASS CONTROL
SET ADDRESS
PARALLEL [ ]
GPIB DIAG on OFF
VOLUME NUMBER
The analyzer is factory-equipped with a remote programming interface using the General Purpose Interface Bus (GPIB). This enables communication between the analyzer and a controlling computer as well as other peripheral devices. This menu indicates the present GPIB controller mode of the analyzer. Three GPIB modes are possible: system controller, talker/listener, and pass control.

When the analyzer is connected to other instruments over GPIB, the GPIB STATUS indicators in the instrument state function block light up to display the current status of the analyzer.
R = remote operation
L = listen mode
T = talk mode
S = service request (SRQ) asserted by the analyzer
The softkey activates the system controller mode. When in this mode, the analyzer can use GPIB to control compatible peripherals, without the use of an external computer. It can output measurement results directly to a compatible printer or plotter, store instrument states using a compatible disk drive, or control a power meter for performing service routines. The power meter calibration function requires system controller or pass control mode.
Talker/Listener Mode
The softkey activates the talker/listener mode, which is the mode of operation most often used. In this mode, a computer controller communicates with the analyzer and other compatible peripherals over the bus. The computer sends commands or instructions to and receives data from the analyzer. All of the capabilities available from the analyzer front panel can be used in this remote operation mode, except for control of the power line switch and some internal tests.
Pass Control Mode
The softkey activates the third mode of GPIB operation: the pass control mode. In an automated system with a computer controller, the controller can pass control of the bus to the analyzer on request from the analyzer. The analyzer is then the controller of the peripherals, and can direct them to plot, print, or store without going through the computer. When the peripheral operation is complete, control is passed back to the computer. Only one controller can be active at a time. The computer remains the system controller, and can regain control at any time.
Preset does not affect the selected controller mode, but cycling the power returns the analyzer to talker/listener mode.
Information on compatible peripherals is provided in the “Options and Accessories” chapter of the reference guide.
Operating Concepts · GPIB Operation

Operating Concepts GPIB Operation
This menu can be accessed by pressing the softkey within the GPIB menu.
In communications through the General Purpose Interface Bus (GPIB), each instrument on the bus is identified by a GPIB address. This decimal-based address code must be different for each instrument on the bus.
This menu lets you set the GPIB address of the analyzer, and enter the addresses of peripheral devices so that the analyzer can communicate with them.
Most of the GPIB addresses are set at the factory and need not be modified for normal system operation. The standard factory-set addresses for instruments that may be part of the system are as follows:
External Disk Drive 00
The address displayed in this menu for each peripheral device must match the address set on the device itself. The analyzer does not have a GPIB switch: its address is set only from the front panel.
These addresses are stored in non-volatile memory and are not affected by preset or by cycling the power.
The instrument’s parallel port can be used in two different modes. By pressing and then toggling the softkey, you can select either the mode or the mode.
Local PARALLEL [ ] [COPY] [GPIO]
The copy mode allows the parallel port to be connected to a printer or plotter for the outputting of test results. To use the parallel port for printing or plotting, you must do the following:

The GPIO mode turns the parallel port into a "general purpose input/output" port.
In this mode the port can be connected to test fixtures, power supplies, and other peripheral equipment that might be used to interact with the analyzer during measurements. This mode is exclusively used in test sequencing.
Operating Concepts · GPIB Operation · The GPIO Mode

Operating Concepts Limit Line Operation
This menu can be accessed by pressing within the system menu.
You can have limit lines drawn on the display to represent upper and lower limits or device specifications with which to compare the test device. Limits are defined in segments, where each segment is a portion of the stimulus span. Each limit segment has an upper and a lower starting limit value. Three types of segments are available: flat line, sloping line, and single point.
Limits can be defined independently for the four channels, up to 22 segments for each channel. These can be in any combination of the three limit types.
Limit testing compares the measured data with the defined limits, and provides pass or fail information for each measured data point. An out-of-limit test condition is indicated in five ways: with a FAIL message on the screen, with a beep, by changing the color of the failing portions of a trace, with an asterisk in tabular listings of data, and with a bit in the GPIB event status register B. (The analyzer also has a BNC rear panel output that includes this status, but is only valid for a single channel measurement.)
Limit lines and limit testing can be used simultaneously or independently. If limit lines are on and limit testing is off, the limit lines are shown on the display for visual comparison and adjustment of the measurement trace. However, no pass/fail information is provided. If limit testing is on and limit lines are off, the specified limits are still valid and the pass/fail status is indicated even though the limit lines are not shown on the display.
Limits are entered in tabular form. Limit lines and limit testing can be either on or off while limits are defined. As new limits are entered, the tabular columns on the display are updated, and the limit lines (if on) are modified to the new definitions. The complete limit set can be offset in either stimulus or amplitude value.
Limits are checked only at the actual measured data points. It is possible for a device to be out of specification without a limit test failure indication if the point density is insufficient. Be sure to specify a high enough number of measurement points in the stimulus menu.
Limit lines are displayed only on Cartesian formats. In polar and Smith chart formats, limit testing of one value is available: the value tested depends on the marker mode and is the magnitude or the first value in a complex pair. The message NO LIMIT LINES DISPLAYED is shown on the display in polar and Smith chart formats.
The list values feature in the copy menu provides tabular listings to the display or a printer for every measured stimulus value. These include limit line or limit test information if these functions are activated. If limit testing is on, an asterisk is listed next to any measured value that is out of limits. If limit lines are on, and other listed data allows sufficient space, the upper limit and lower limit are listed, together with the margin by which the device data passes or fails the nearest limit.

Operating Concepts Limit Line Operation
If limit lines are on, they are plotted with the data on a plot. If limit testing is on, the PASS or FAIL message is plotted, and the failing portions of the trace that are a different color on the display are also a different color on the plot. If limits are specified, they are saved in memory with an instrument state.
This menu allows you to specify limits for limit lines or limit testing, and presents a table of limit values on the display. Limits are defined in segments. Each segment is a portion of the stimulus span. Up to 22 limit segments can be specified for each channel. The limit segments do not have to be entered in any particular order: the analyzer automatically sorts them and lists them on the display in increasing order of start stimulus value.
For each segment, the table lists the segment number, the starting stimulus value, upper limit, lower limit, and limit type. The ending stimulus value is the start value of the next segment, or a segment can be terminated with a single point segment. You can enter limit values as upper and lower limits or delta limits and middle value. As new limit segments are defined, the tabular listing is updated. If limit lines are switched on, they are shown on the display.
If no limits have been defined, the table of limit values shows the notation EMPTY. Limit segments are added to the table using the softkey or edited with the softkey, as previously described. The last segment on the list is followed by the notation END.
Edit Segment Menu
This menu sets the values of the individual limit segments. The segment to be modified, or a default segment, is selected in the edit limits menu. The stimulus value can be set with the controls in the entry block or with a marker (the marker is activated automatically when this menu is presented). The limit values can be defined as upper and lower limits, or delta limits and middle value. Both an upper limit and a lower limit (or delta limits) must be defined: if only one limit is required for a particular measurement, force the other out of range (for example +500 dB or −500 dB).
As new values are entered, the tabular listing of limit values is updated.
Segments do not have to be listed in any particular order: the analyzer sorts them automatically in increasing order of start stimulus value when the key in the edit limits menu is pressed. However, the easiest way to enter a set of limits is to start with the lowest stimulus value and define the segments from left to right of the display, with limit lines turned on as a visual check.
Phase limit values can be specified between +500° and −500°. Limit values above +180° and below −180° are mapped into the range of −180° to +180° to correspond with the range of phase data values.
This menu allows the complete limit set to be offset in either stimulus value or amplitude value. This is useful for changing the limits to correspond with a change in the test setup, or for device specifications that differ in stimulus or amplitude. It can also be used to move the limit lines away from the data trace temporarily for visual examination of trace detail.

Operating Concepts Knowing the Instrument Modes
There are three major instrument modes of the analyzer:
This is the standard mode of operation for the analyzer, and is active after you press
or switch on the AC power. This mode uses the analyzer's internal source.
Pressing returns the analyzer to the "normal" network analyzer operating mode.
System INSTRUMENT MODE NETWORK ANALYZER
In the tuned receiver mode, the analyzer's receiver operates independently of any signal source. This mode is not phase-locked and functions in all sweep types. The analyzer tunes the receiver to a synthesized CW input signal at a precisely specified frequency. All phase lock routines are bypassed, increasing sweep speed significantly. The external source must be synthesized, and must drive the analyzer's external frequency reference. The analyzer's internal source frequency is not accurate, and the internal source should not be used in the tuned receiver mode.
Using the analyzer's tuned receiver mode is useful for automated test applications where an external synthesized source is available and applications where speed is important. Although the tuned receiver mode can function in all sweep types, it is typically used in CW applications.
System INSTRUMENT MODE TUNED RECEIVER
Preset · Typical test setup

Operating Concepts Knowing the Instrument Modes
Figure 7-46 Typical Test Setup for Tuned Receiver Mode
Tuned Receiver Mode In-Depth Description
If you press , the analyzer receiver operates independently of any signal source.
System INSTRUMENT MODE TUNED RECEIVER
The following features and limitations apply to the tuned receiver mode:
For more information on using the tuned receiver mode, refer to “Tuned Receiver Mode” on page 2-27.
Refer to “Conversion Loss Using the Frequency Offset Mode” on page 2-13 for information on frequency offset operation.

Operating Concepts Knowing the Instrument Modes


Safety and Regulatory Information General Information
Clean the cabinet, using a dry or damp cloth only.
Product maintenance agreements and other customer assistance agreements are available for Agilent Technologies products.
For any assistance, contact your nearest Agilent Technologies Sales and Service Office.
If you are sending the instrument to Agilent Technologies for service, ship the analyzer to the nearest service center for repair, including a description of any failed test and any error message. Ship the analyzer using the original or comparable antistatic packaging materials.

Safety and Regulatory Information
Table 8-1 Contacting Agilent
United States (tel) 1 800 452 4844
Latin America (tel) (305) 269 7500 (fax) (305) 269 7599
Canada (tel) 1 877 894 4414 (fax) (905) 282-6495
Europe (tel) (+31) 20 547 2323 (fax) (+31) 20 547 2390
New Zealand (tel) 0 800 738 378 (fax) (+64) 4 495 8950
Japan (tel) (+81) 426 56 7832 (fax) (+81) 426 56 7840
Australia (tel) 1 800 629 485 (fax) (+61) 3 9210 5947
Asia Call Center Numbers
Country Phone Number Fax Number
Singapore 1-800-375-8100 (65) 836-0252
Philippines (632) 8426802 1-800-16510170 (PLDT Subscriber Only)
(632) 8426809 1-800-16510288 (PLDT Subscriber Only)
Thailand (088) 226-008 (outside Bangkok) (662) 661-3999 (within Bangkok)
Hong Kong 800-930-871 (852) 2506 9233
Taiwan 0800-047-866 (886) 2 25456723
People’s Republic of China
800-810-0189 (preferred) 10800-650-0021 10800-650-0121

Safety and Regulatory Information Safety Symbols
The following safety symbols are used throughout this manual. Familiarize yourself with each of the symbols and its meaning before operating this instrument.
The instruction documentation symbol. The product is marked with this symbol when it is necessary for the user to refer to the instructions in the documentation.
The CE mark is a registered trademark of the European Community. (If accompanied by a year, it is when the design was proven.)
The CSA mark is a registered trademark of the Canadian Standards Association.
This is a symbol of an Industrial Scientific and Medical Group 1 Class A product.
The C-Tick mark is a registered trademark of the Australian Spectrum Management Agency.
ICES / NMB-001 This is a marking to indicate product compliance with the Canadian Interference-Causing Equipment Standard (ICES-001).

Safety and Regulatory Information
CAUTION Install the instrument so that the ON/OFF switch is readily identifiable and is easily reached by the operator. The ON/OFF switch or the detachable power cord is the instrument disconnecting device. It disconnects the mains circuits from the mains supply before other parts of the instrument. Alternately, an externally installed switch or circuit breaker (which is readily identifiable and is easily reached by the operator) may be used as a disconnecting device.
Safety Considerations

Safety and Regulatory Information Safety Considerations

Safety and Regulatory Information
CAUTION VENTILATION REQUIREMENTS: When installing the product in a cabinet, the convection into and out of the product must not be restricted. The ambient temperature (outside the cabinet) must be less than the maximum operating temperature of the product by 4° C for every 100 watts dissipated in the cabinet. If the total power dissipated in the cabinet is greater that 800 watts, then forced convection must be used.
Safety Considerations

Safety and Regulatory Information Safety Considerations
This network analyzer complies with German FTZ 526/527 Radiated Emissions and Conducted Emission requirements.
This is to declare that this instrument is in conformance with the German Regulation on Noise Declaration for Machines (Laermangabe nach der Maschinenlaermrerordung −3. GSGV Deutschland).
LpA
Operator Position am Arbeitsplatz
Normal Operation normaler Betrieb
per ISO 7779 nach DIN 45635 t. 19
This ISM device complies with Canadian ICE-001.
Cet appareil ISM est conforme a la norme NMB du Canada.
Acoustic Noise Emission/Geraeuschemission

Safety and Regulatory Information

Declaration of Conformity

Safety and Regulatory Information Declaration of Conformity

modifying, 1-101 commands that require clean
data trace, 1-19
calibration, 7-55 calibration standards, 6-5 calibration techniques
saving to display memory, 1-19 decision making functions, 1-112 decoupled
sweep, 1-106 commands that sequencing
improper, 5-4 calibration, measurement, 7-37 calibration, receiver, 6-15 calibration, TRL*/LRM*, 7-67 calling the next measurement
channel power, 1-13 stimulus, 1-13 decreasing
completes before next command, 1-105 comma-separated values. See
frequency span, 5-11 sweep rate, 5-9 time delay, 5-9 decrementing the loop counter,
CSV format compensating for directional
sequence, 2-30 capabilities
coupler response, 6-35 compliance with German noise
mixer measurement, 2-3 capacitance, fringe, 6-6 cause of measurement problems,
requirements, 8-8 computer, what you can save, 4-37 confidence check, 6-65 configuring
2-32 default colors, 1-22 defining
plot function, 4-15 print function, 4-6 defining a print function
5-8 center frequency, setting, 1-35 changing
plot function, 4-10 configuring two sources, 2-30 connecting
sequence title, 1-103 system bandwidth, 5-16 changing the ripple limits color,
color printer, using, 4-6 resetting print parameters to
device under test, 1-4 required test equipment, 1-4 connector
default value, 4-7 delay block, electrical, 7-7 delay, determining electrical, 6-73 delay, electrical, 7-33 delay, group, 1-46, 2-35 deleting
1-88 channel coupling, 7-10 channel position softkey, 1-17 channel stimulus coupling, 7-14 characteristics of the filter, 1-68 characterizing a duplexer, 1-49
care, 5-3 repeatability, 5-4 continuous correction mode,
using, 6-38 continuous markers, 1-24 conversion, 7-7 conversion compression using
commands, 1-100 frequency signals, 6-35 line segments, 1-78 deleting a file, 4-52
definitions, 1-49 procedure, 1-49 characterizing microwave
frequency offset mode, 2-40 conversion loss, 2-20 conversion loss using frequency
systematic errors, 7-41 device measurement, 7-46 one-port error model, 7-41 two-port error model, 7-47 choosing
all files, 4-52 instrument state file, 4-52 delta markers, 1-28 demodulating the results of the
offset mode, 2-13 conversion menu, 7-22 copy mode, 7-80 correction, sampler/IF, 7-6 coupling display markers, 1-31 coupling, channel stimulus, 7-14 creating
forward transform, 3-23 designing your own fixture, 6-49 detecting IF delay, 5-11 detection, IF, 7-6 deviation from linear phase, 1-46 device measurement, 7-46 device measurements, 6-4 device under test
display elements, 4-15 measurement parameters, 1-4 plot speed, 4-18 scale, 4-17 chop sweep mode, activating, 5-13 CITIfile, 4-41 clarifying type-N connector sex,
flat limit lines, 1-73 sequence, 1-98 sloping limit line, 1-75 user-defined TRL calibration
measuring, 1-5 device under test, connecting, 1-4 device, bilateral, 6-22, 6-25 device, noninsertable, 6-69 direct sampler access
6-4 clearing a sequence from memory,
kit, 6-50 user-defined TRM calibration
kit, 6-54 creating single point limits, 1-77 crosstalk, 7-40 CSV format, saving measurement
changing the ripple limits, 1-88 ripple test message, 1-87 color printer, using, 4-6 colors of the display, adjusting,
configurations, using, 5-17 directional coupler response,
results in, 4-44 CW frequency, setting, 1-38 CW time measurements, 3-22 CW time sweep, 7-19
compensating for, 6-35 discrete markers, 1-24 disk
1-22 default, 1-22 intensity, 1-22 modified, 1-23 modify colors menu, 1-22 command
formatting, 4-54 disk, plotting a measurement to,
D data formats, ASCII, 4-41 data processing, 7-5
4-13 display elements, choosing, 4-15 display functions, 1-10
deleting, 1-100 inserting, 1-101
active channel display, 1-11
processing details, 7-6
TRL*/LRM* two-port · 1-102 color · Index-2

titling, 1-11 adjusting colors of the display,
dynamic range, increasing, 5-15
error-correction, vector, 7-7 errors, measurement, 7-38 exit HPGL mode, 4-26
1-22 blanking the display, 1-21 data trace
E ECal, 6-58–6-75
sending to the printer, 4-26 external
adapter removal calibration,
saving to display memory,
calibration, 5-13 disk drive, 4-55 R channel inputs, 2-10
6-69–6-75 calibration, 6-58–6-68 confidence check, 6-65 equipment, 6-59 isolation calibration, 6-61 manual thru, 6-60 module information, 6-64 options, 6-60 service menu, 6-67 edit limits menu, 7-83 edit segment menu, 7-83 editing a sequence, 1-100
4 Param Displays softkey, 1-18 Channel Position softkey, 1-17 customizing, 1-17 viewing, 1-14 measurement data
F fabricating and defining
calibration standards for TRL/LRM, 7-74 fast 2-port calibration, 5-13 fault location measurements
dividing by the memory trace,
using low pass, 3-18 features that operate differently
1-20 subtracting memory trace,
1-20 viewing, 1-20 memory math functions, 1-19 memory trace
deleting commands, 1-100 editing line segments, 1-78 editing ripple limits, 1-84–1-86 electrical
in sequence, 1-105 feedthrough, RF, 2-47 file
deleting, 4-52 index numbers, 4-45 recalling, 4-54 renaming, 4-53 sequential CSV naming of, 4-45 to delete all, 4-52 filter, characteristics, 1-68 finite impulse width (or rise time),
viewing, 1-20 memory traces, 1-19 primary measurement
length, 1-43 offset, 6-6 electrical delay, 7-33 electrical delay block, 7-7 electrical delay, determining, 6-73 electrical delay, setting, 1-37 electrically long devices, 5-4 electronic calibration. See ECal eliminating unwanted mixing and
dual channel mode with
decoupled channel power, 1-13 dual channel mode with
decoupled stimulus, 1-13 primary measurement
3-27 fixed IF mixer adjustments
channels, 1-12 viewing, 1-12 ratio measurements in channel
leakage signals, 2-6 embedding loop counter value in
tuned receiver mode, 2-27 fixed IF mixer measurements,
title, 1-106 enhanced frequency response
1 and 2, to, 1-20 display intensity, 1-22 display markers
2-27 addressing and configuring two
error correction, 6-22 enhanced reflection error
sources, 2-30 calling the next measurement
coupling, 1-31 uncoupling, 1-31 display markers, activating, 1-25 display memory, 1-19, 7-8 display reference value, setting,
correction, 6-22, 6-25 enhancement, accuracy, 7-37 entering the power sensor
sequence, 2-30 decrementing the loop counter,
calibration data, 6-34 deleting frequency signals, 6-35 editing frequency segments,
2-32 frequency list sweep of 26
points, 2-29 incrementing the source
1-37 display ripple test limits, 1-87 display ripple test values, 1-88 displayed measurement, titling,
frequencies, 2-32 initializing loop counter value to
26, 2-30 labeling the screen, 2-32 presetting the instrument, 2-28 prompting user to connect mixer
adapters, 6-47 error correcting measurements,
minimizing error while using
4-32 displaying the bandwidth
markers, 1-95 displaying the bandwidth value,
test setup, 2-29 response calibration, 2-29 sequence 1 setup, 2-27 sequence 2 setup, 2-32 taking data, 2-32 tuned receiver mode, 2-29 fixed markers, 1-29 fixtures, 6-48
1-96 drift, frequency, 5-5 drift, temperature, 5-5 dual channel mode with decoupled
enhanced frequency response,
6-22 enhanced reflection, 6-22, 6-25 frequency response, 6-12 full two-port, 6-29 one-port reflection, 6-26 error-correction, 1-5
power, 1-13 dual channel mode with decoupled
stimulus, 1-13 duplexer, characterizing, 1-49
stimulus state, 6-9
designing your own fixture, 6-49

flat limit lines, 1-73 floppy disk, what you can save,
swept RF/IF conversion loss,
6-17 frequency response error
2-20 high dynamic range
4-37 form feed sequence, 4-26
corrections, 6-12 receiver calibration, 6-15 response error correction for
measurement, 2-24 high dynamic range swept RF/IF
sending to the printer, 4-26 format, 7-8 format arrays, 7-8 format markers, polar, 1-32 formats, analyzer display, 7-24 formatting a disk, 4-54 forward stepping in edit mode,
conversion loss measurement parameters for IF
reflection measurements, 6-12 response error correction for
range, 2-20 power meter calibration over IF
range, 2-20 power meter calibration over RF
transmission measurements, 6-14 frequency response of calibration
range, 2-23 receiver calibration over IF
standards, 6-6 electrical offset, 6-6 fringe capacitance, 6-6 fringe capacitance, 6-6 full two-port calibration, 7-55 full two-port error correction, 6-29
measurements, 3-22 demodulating the results of the
range, 2-22 RF frequency range, 2-22 high power
forward transform, 3-23 forward transform range, 3-24 interpreting the forward
measurements, 1-59, 1-65 high power measurements
transform horizontal axis, 3-23 interpreting the forward
additional setup, 1-60 attenuator settings, selecting,
G gain, 1-57 gain compression
1-61 final setup, 1-63 initial setup, 1-59 power levels, determining, 1-59 power ranges, selecting, 1-61 horizontal axis, 3-13, 3-14, 3-16,
transform vertical axis, 3-22 forward transform mode, 3-4 forward transform range, 3-24 four-channel display
measuring, 1-53 using linear sweep, 1-55 gating, 3-35, 7-7
4 Param Displays softkey, 1-18 Channel Position softkey, 1-17 customizing, 1-17 viewing, 1-14 Freelance, using, 4-23 frequency
selecting gate shape, 3-36 setting the gate, 3-35 general, 8-7 generating files in a loop counter
3-20, 3-23 how RF and IF are defined, 2-7 how to
example, 1-116 German FTZ emissions
measurements, 1-59 HPGL
segments, editing, 6-34 signals, deleting, 6-35 span, decreasing, 5-11 span, setting, 1-36 frequency bands
requirements, 8-8 gosub sequence command, 1-107 GPIB operation
compatible printer, 4-21 initialization sequence, sending
to the printer, 4-26 initialization sequence, storing,
address menu, 7-80 GPIB STATUS indicators, 7-79 local key, 7-78 pass control mode, 7-79 system controller mode, 7-79 talker/listener mode, 7-79 using the parallel port, 7-80 GPIO mode, 1-107, 7-81 graphic files
adding, 1-85 changing, 1-84 deleting, 1-86 setting, 1-83 frequency drift, 5-5 frequency list sweep of 26 points,
4-25 HPGL compatible printer, 4-25 HPGL/2 compatible printer, 4-10
I IF bandwidth reduction, 7-35 IF bandwidth, setting segment,
2-29 frequency offset mode operation,
saving measurement results as,
7-19 IF delay, detecting, 5-11 IF detection, 7-6 IF range
2-9 frequency offset mode, conversion
compression, 2-40 frequency offset mode, conversion
moving marker information off
the grids, 1-26 group delay, 1-46 group delay format, 7-25 group delay measurements, 2-35 group delay principles, 7-29
measurement parameters, 2-20 power meter calibration, 2-20 receiver calibration, 2-22 IF, defining, 2-7 imaginary format, 7-29 improving measurement results,
loss, 2-13 frequency offset operation, 7-85 frequency range, setting, 1-5 frequency range, setting for time
domain low pass, 3-15 frequency response, 7-41 frequency response and isolation
H high dynamic range
5-8 decreasing the sweep rate, 5-9 decreasing time delay, 5-9
error corrections, 6-17 reflection measurements, 6-19

forward stepping in edit mode,
forward transform vertical axis,
improving raw source match and
load match for TRL*/LRM* calibration, 7-71 increase sweep speed
1-106 gosub sequence command, 1-107 GPIO mode, 1-107 limit test decision making,
3-22 low pass response horizontal
axis, 3-16 low pass response vertical axis,
using fast 2-port calibration,
5-13 increasing dynamic range, 5-15
1-112 loop counter decision making,
3-16 low pass step transmission
increasing test port input power,
response horizontal axis, 3-20 low pass step transmission
1-113 sequence decision making
5-15 reducing receiver crosstalk, 5-15 reducing the receiver noise floor,
menu, 1-112 sequence size, 1-106 sequence that jumps to itself,
response vertical axis, 3-20 introduction to time domain
measurements, 3-3 isolation, 7-40, 7-69
accuracy, 5-4 connector repeatability, 5-4 frequency drift, 5-5 interconnecting cables, 5-4 performance verification, 5-5 reference plane and port
1-112 sequencing special functions
menu, 1-112 titles, 1-106 TTL I/O menu, 1-108 TTL input decision making,
averaging, 6-61 calibrating using ECal, 6-61 calibration, omitting, 6-4 error corrections and frequency
1-108, 1-112 TTL out menu, 1-112 TTL output for controlling
response, 6-17 isolation example measurements,
extensions, 5-5 temperature drift, 5-5 increasing sweep speed, 5-10
2-45 LO to RF isolation, 2-45 RF feedthrough, 2-47 SWR/return loss, 2-50
peripherals, 1-108 indicators, GPIB STATUS, 7-79 initializing loop counter value to
activating chop sweep mode,
5-13 decreasing the frequency span,
26, 2-30 input ports menu, 7-23 inserting a command, 1-101 insertion phase response, 1-7, 1-8 instrument markings, 8-4 instrument modes, 7-84
5-11 reducing the averaging factor,
J jpeg files, saving results as, 4-46
5-12 reducing the number of
measurement points, 5-12 setting the auto sweep time
K knowing the instrument modes,
frequency offset operation, 7-85 network analyzer mode, 7-84 tuned receiver mode, 7-84 instrument state
mode, 5-12 setting the sweep type, 5-12 using external calibration, 5-13 using swept list mode, 5-10 viewing a single measurement
file, deleting, 4-52 files, 4-47 re-saving, 4-51 saving, 4-38, 7-65 saving and recalling, 4-36 interconnecting cables, 5-4 internal memory, 4-36 internal R channel inputs, 2-10 interpolated error correction, 6-8 interpolation in power meter
L labeling the screen, 2-32 leakage signals, eliminating
channel, 5-13 widening the system bandwidth,
unwanted, 2-6 limit line operation, 7-82
edit limits menu, 7-83 edit segment menu, 7-83 offset limits menu, 7-83 limit lines
5-12 increasing test port input power,
5-15 incrementing the source
frequencies, 2-32 in-depth sequencing information,
creating flat limit lines, 1-73 creating single point limits, 1-77 editing line segments, 1-78 measurement parameters, 1-72 offsetting limit lines, 1-79 running a limit test, 1-79 sloping limit line, 1-75 using to test a device, 1-72 limit test
calibration, 6-34 interpreting
1-105 autostarting sequences, 1-106 commands completed before
next sequence begins, 1-105 commands that require a clean
horizontal axis, 3-13 bandpass reflection response
sweep, 1-106 decision making functions,
vertical axis, 3-13 bandpass transmission response
horizontal axis, 3-14 bandpass transmission response
decision making, 1-112 example sequence, 1-118 limit test, running, 1-79
1-112 embedding loop counter value in
title, 1-106 features that operate differently
vertical axis, 3-14 forward transform horizontal
activating the limit test, 1-79 reviewing the limit line
in a sequence, 1-105

line segments, editing, 1-78
deleting line segments, 1-78 line types, selecting, 4-17 linear frequency sweep, 7-15 linear magnitude format, 7-27 linear phase, deviation, 1-46 linear sweep, 1-55 linearity, phase, 2-35 list values, printing or plotting,
measurements, 3-5 manual mode, 7-9 manual sweep time mode, 7-11 manual thru, 6-60 margin
systematic errors, 7-41 measurement errors, 7-38 measurement considerations, 2-4
and leakage signals, 2-6 frequency offset mode operation,
ripple test value, 1-88, 1-91 markers, 1-24
2-9 how RF and IF are defined, 2-7 internal and external R channel
measurement data, 1-42 continuous, 1-24 coupling display markers, 1-31 CW frequency, setting, 1-38 delta markers, 1-28 discrete, 1-24 display markers
4-34 LO frequency accuracy and
inputs, 2-10 LO frequency accuracy and
stability, 2-10 LO to RF isolation, 2-45 load match, 7-40, 7-71 load mismatches, minimizing, 2-4 loading a sequence from a disk,
stability, 2-10 minimizing source and load
mismatches, 2-4 power meter calibration, 2-12 reducing the effect of spurious
activating, 1-25 fixed marker
1-104 local key, 7-78 log magnitude format, 7-24 logarithmic frequency sweep, 7-15 loop counter
responses, 2-5 measurement data, 1-20
activating, 1-29 moving marker information off
dividing by the memory trace,
of the grids, 1-26 polar format markers, 1-32 setting measurement
1-20 viewing, 1-20 measurement data trace, 1-20
decision making, 1-113 example sequence, 1-115 value, 2-30 loss of power meter calibration
parameters, 1-34 smith chart markers, 1-33 specific amplitude, searching
subtracting memory trace, 1-20 measurement error
data, 6-33 low pass impulse mode, 3-20 lower stopband parameters, 1-69
crosstalk, 7-40 frequency response, 7-41 isolation, 7-40 load match, 7-40 measurement errors
for, 1-39 uncoupling display markers,
1-31 masking, 3-26 matched adapters, 6-45 maximum amplitude, searching
directivity, 7-38 source match, 7-39 measurement parameters, 1-68,
measuring, 1-7 measuring response, 1-7 maintaining testport output
for, 1-39 maximum and minimum, 1-94 maximum bandwidth, 1-94 measurement
6-4 center frequency, setting, 1-35 choosing, 1-4 display reference value, setting,
power during sweep retrace, 5-7 maintenance, 8-2 making a basic measurement, 1-4
accuracy, increasing, 5-4 calibration, power meter, 6-33 fault location using low pass,
1-37 electrical delay, setting, 1-37 for IF range, 2-20 frequency span, setting, 1-36 lower stopband parameters,
parameters, 1-4 connecting required test
3-18 high dynamic range, 2-24 isolation example, 2-45 non-coaxial, 6-48 plotting to a disk, 4-13 printing or plotting results, 4-3 reflection response, 3-9 results, outputting, 1-6 results, saving, 4-39 results, saving graphically, 4-46 setting, 1-5 time domain, 3-3 transmission measurements in
equipment, 1-4 connecting the device under
1-69 markers, setting with, 1-34 passband parameters, 1-69 start frequency, setting, 1-34 stop frequency, setting, 1-35 upper stopband parameters,
test, 1-4 error-correction, 1-5 frequency range, setting, 1-5 measurement, setting, 1-5 measuring the device under
test, 1-5 outputting measurement
basic, 1-4 high power, 1-59, 1-65 phase or group delay, 2-35 measuring
results, 1-6 source power, setting, 1-5 making non-coaxial
time domain low pass, 3-19 transmission response, 3-5 measurement calibration, 7-37
measurements, 6-48 fixtures, 6-48 making reflection response
device under test, 1-5 gain and reverse isolation
accuracy enhancement, 7-37,

insertion phase response, 1-7,
noise reduction techniques, 7-34
for, 1-39 minimum bandwidth, 1-94 minimum sweep time, 7-11 mixer
averaging, 7-34 IF bandwidth reduction, 7-35 smoothing, 7-35 non-coaxial
1-8 separate transmission paths
through the test device using low pass impulse mode, 3-20 small signal transient response
fixed IF measurements, 2-27 measurement, 2-3 measurement diagram, using,
making measurements, 6-48 non-coaxial devices, calibrating
using low pass step, 3-19 measuring amplifiers, 1-52
for, 6-50 noninsertable device, 6-69 noninsertable devices, calibrating
2-17, 2-23 mixing signals, eliminating
high power measurements, 1-65 high power measurements
unwanted, 2-6 mode
making, 1-59 measuring gain and reverse
auto sweep time, 5-12, 7-11 chop sweep, 5-13 continuous correction, 6-38 copy, 7-80 frequency offset, 2-40 GPIO, 7-81 low pass impulse, 3-20 manual sweep time, 7-11 network analyzer, 7-84 pass control, 7-79 system controller, 7-79 talker/listener, 7-79 time domain bandpass, 3-12 time domain low pass, 3-15 tuned receiver, 7-84 model
isolation simultaneously, 1-57 measuring gain compression,
O offset and scale, 7-8 offset limits menu, 7-83 offset, electrical, 6-6 offsetting limit lines, 1-79 omitting isolation calibration, 6-4 one-port calibration, S11 and S22,
1-53 measuring electrical length, 1-43 measuring gain compression, 1-53
linear sweep, 1-55 using linear sweep, 1-55 measuring magnitude, 1-7
7-55 one-port error model, 7-41 one-port reflection error
magnitude response, 1-7 measuring phase distortion, 1-43,
correction, 6-26 operating parameters, printing or
plotting, 4-34 operation, frequency offset, 7-85 operation, GPIB, 7-78 operation, limit line, 7-82 operation, system, 7-3 operation, trace math, 7-7 output power, 7-9
1-45 deviation from linear phase,
1-46 group delay, 1-46 memory math functions, 1-19 memory trace, 1-19, 1-20
one-port error, 7-41 two-port error, 7-47 modified colors, 1-23
viewing, 1-20 memory, display, 7-8 menu
recalling, 1-23 saving, 1-23 modify colors menu, 1-22 modifying
power coupling options, 7-10 understanding power ranges,
address, 7-80 analog in, 7-22 calibration kit, 7-58 conversion, 7-22 edit limits, 7-83 edit segment, 7-83 input ports menu, 7-23 offset limits, 7-83 segment, 7-16 S-parameter, 7-22 stepped edit list, 7-16 stepped edit subsweep, 7-16 swept edit list, 7-17 swept edit subsweep, 7-17 microprocessor, 7-4 microwave connector care, 5-3 microwave systematic errors,
cal kit through definition, 6-46 command, 1-101 standard definitions, 6-50 modifying calibration kits, 7-57
measurement results, 1-6 multiple plots to a single page
using a plotter, 4-27 plot files, 4-14 plot files from a PC to a plotter,
calibration kit menu, 7-58 saving to a disk, 7-65 verifying performance, 7-65 module information, 6-64 multiple
4-24 single page plots using a printer,
4-26 outputting plot files from a PC to
an HPGL compatible printer, 4-25 sending the exit HPGL mode
plotting from a disk, 4-28 measurements, plotting on a full
page, 4-29 plots, outputting to a single page
and form feed sequence to the printer, 4-26 sending the HPGL initialization
using a plotter, 4-27
characterizing, 7-41 minimizing
sequence to the printer, 4-26 sending the plot file to the
error when using adapters, 6-47 source and load mismatches, 2-4 minimum allowable stop
N N dB Point, 1-94 names for CSV files, 4-45 naming files generated by a
printer, 4-26 storing the exit HPGL mode and
sequence, 1-103 network analyzer mode, 7-84
form feed sequence, 4-26

choosing display elements, 4-15 choosing plot speed, 4-18 choosing scale, 4-17 resetting plotting parameters to
storing the HPGL initialization
correction mode, 6-36 power ranges, 7-9
automatic mode, 7-9 manual mode, 7-9 power ranges, selecting, 1-61 power sensor calibration data,
P page quadrants, plotting
default values, 4-18 selecting auto-feed, 4-15 selecting line types, 4-17 selecting pen numbers and
measurements in, 4-30 parameters
entering, 6-34 power sweep, 7-19 power, output, 7-9 primary measurement channels,
lower stopband, 1-69 measurement, 1-4, 1-68, 6-4 passband, 1-69 upper stopband, 1-69 pass control mode, 7-79 passband parameters, 1-69 PC interface unit, 6-59 PC, to view files on, 4-22 pen numbers and colors, selecting,
colors, 4-16 plot speed, choosing, 4-18 plotting
measurement results, 4-3 measurement to a disk, 4-13 measurements in page
viewing, 1-12 principles, group delay, 7-29 print
quadrants, 4-30 multiple measurements on a full
aborting a process, 4-33 print function
page, 4-29 multiple measurements per
configuring, 4-4 defining, 4-6 printer
4-16 pen plotter, 4-19, 4-20
plotting to, 4-12 performance verification, 5-5 performance, verifying, 7-65 performing
page from a disk, 4-28 multiple measurements per
color printer, using, 4-6 HPGL compatible printer, 4-21,
page using a pen plotter, 4-20 one measurements per page
4-25 HPGL/2 compatible printer,
2-port error corrections, 6-42, 6-71 TRL calibration, 6-52 TRM calibration, 6-56 peripheral equipment, required,
using a pen plotter, 4-19 parameters, resetting to default
values, 4-18 plotting to an HPGL compatible
measurement results, 4-3 multiple measurements per
printer, 4-21 solving problems, 4-35 plotting a measurement to a disk
page, 4-9 one measurement per page, 4-8 parameters, resetting to default
7-4 per-raw data arrays, 7-7 phase
linearity, 2-35 measurements, 2-35 tracking, 2-39 phase distortion, 1-43, 1-45 phase format, 7-24 phase or group delay
to output the plot files, 4-14 polar format, 7-27 polar format markers, 1-32 port extensions, 5-5 power coupling options, 7-10
values, 4-7 sequence, 1-105 solving problems, 4-35 printing or plotting the list values
or operating parameters, 4-34 entire list of values, 4-34 single page of values, 4-34 procedure
channel coupling, 7-10 test port coupling, 7-10 power levels, determining, 1-59 power meter
measurements, 2-35 phase linearity and group delay,
high power measurements, 1-59 procedure, TRL calibration, 7-72 procedures for error correcting
2-35 phase measurements, 2-35 places where you can save, 4-36 plot
calibration, 2-12 calibration over IF range, 2-20 calibration over RF range, 2-23 power meter measurement
measurements, 6-10 types of error correction, 6-10 processing details, 7-6
aborting a process, 4-33 plot files
calibration, 6-33 calibrating the analyzer receiver
accuracy enhancement, 7-7 ADC, 7-6 conversion, 7-7 display memory, 7-8 electrical delay block, 7-7 format, 7-8 format arrays, 7-8 gating, 7-7 IF detection, 7-6 pre-raw data arrays, 7-7 ratio calculations, 7-6 raw arrays, 7-7 sampler/IF correction, 7-6
outputting from a PC to a
plotter, 4-24 outputting from a PC to an
to measure absolute power, 6-39 compensating for directional
HPGL compatible printer, 4-25 sending to the printer, 4-26 to output, 4-14 to view on a PC, 4-22 plot function, configuring, 4-10
coupler response, 6-35 entering the power sensor
calibration data, 6-34 interpolation in power meter
calibration, 6-34 loss of power meter calibration
plotting to a pen plotter, 4-12 plotting to an HPGL/2
data, 6-33 using continuous correction
compatible printer, 4-10 plot function, defining, 4-15

scale and offset, 7-8 smoothing, 7-8 sweep-to-sweep averaging, 7-6 trace math operation, 7-7 transform, 7-8 vector error-correction, 7-7 processing, data, 7-5 prompting user to connect mixer
using the calculation, 2-22 using the mixer measurement
bandpass interpreting the band pass
diagram, 2-17, 2-23 RF range
reflection response vertical axis, 3-13 reflection measurements using
power meter calibration, 2-23 RF, defining, 2-7 ripple limit testing, 1-81–1-91 ripple limits
bandpass mode, 3-12 interpreting the bandpass
test setup, 2-29 purging a sequence from a disk,
editing, 1-84–1-86 running the test, 1-86–1-91 setting, 1-81–1-84 ripple test
reflection response horizontal axis, 3-13 reflection response
measurements, making, 3-9 relative velocity factor, adjusting,
absolute value, 1-89 displaying limits, 1-87 displaying values, 1-88 frequency bands, 1-83 margin value, 1-91 message color, 1-87 starting and stopping, 1-86 running a bandwidth test,
R R channel inputs, internal and
3-12 removing adapter delay, 6-42 removing the effect of an adapter,
external, 2-10 range, 3-30
forward transform, 3-24 resolution, 3-34 ratio calculations, 7-6 ratio measurements in channel 1
6-74 renaming a file, 4-53 repeatability, connector, 5-4 repetitive switching of the
attenuator, 7-13 required peripheral equipment,
1-94–1-97 running a limit test, 1-79 running a sequence, 1-100 running the ripple limits test,
and 2, to, 1-20 raw arrays, 7-7 raw source match, 7-71 real format, 7-29 recall time, reducing, 5-18 recalling a file, 4-54
7-4 required test equipment,
connecting, 1-4 requirements for TRL standards,
solving problems, 4-55 recalling instrument states, 4-36 receiver
7-72 re-saving an instrument state,
S S11 and S22 one-port calibration,
7-55 S2P data format, 4-42 safety considerations, 8-5
calibration, 6-15 crosstalk, reducing, 5-15, 5-18 noise floor, reducing, 5-15 receiver block, 7-4 reducing
plotting parameters to default
values, 4-18 printing parameters to default
before applying power, 8-5 compliance with German FTZ
values, 4-7 resolution, 3-32
emissions requirements, 8-8 compliance with German noise
averaging factor, 5-12 effect of spurious responses, 2-5 number of measurement points,
range resolution, 3-34 response resolution, 3-32 response
requirements, 8-8 general, 8-7 safety earth ground, 8-5 servicing, 8-6 safety earth ground, 8-5 safety symbols, 8-4 sample-and-sweep correction
5-12 reducing recall time, 5-18 reducing trace noise, 5-16
calibration, 2-29 error correction for reflection
measurements, 6-12 error correction for transmission
activating averaging, 5-16 changing system bandwidth,
measurements, 6-14 magnitude, 1-7 resolution, 3-32 response and isolation
mode, using, 6-36 sampler/IF correction, 7-6 saving
5-16 reduction, IF bandwidth, 7-35 reference plane and port
extensions, 5-5 reflection measurements
calibration data, 6-5, 7-65 data trace, 1-19 instrument state, 4-38, 7-65 measurement results, 4-39 measurement results
calibration, 7-54 response and isolation error
correction, 6-19 response error correction, 6-12 reflection measurements in time
response and isolation error
correction for reflection measurements, 6-19 response calibration, 7-54 restarting a calibration, 6-5 reverse isolation, 1-57 reviewing the limit line segments,
graphically, 4-46 modified calibration kits, 7-65 to a disk, 7-65 saving a file
domain low pass, 3-16 interpreting the low pass
response horizontal axis, 3-16 interpreting the low pass
solving problems, 4-55 saving and recalling instrument
1-79 RF feedthrough, 2-47 RF frequency range, 2-22
response vertical axis, 3-16

places where you can save, 4-36 what you can save to a
tracking the amplitude, 1-41 spreadsheet, saving test file for a,
low pass, 3-15 gate, 3-35 setting ripple limits, 1-81–1-84 setting segment IF bandwidth,
frequency range for time domain
computer, 4-37 what you can save to a floppy
4-44 spurious responses, reducing the
disk, 4-37 what you can save to the
effect of, 2-5 standards, calibration, 6-5 start frequency, setting, 1-34 starting the ripple test, 1-86 statistics of measurement data,
7-19 setting segment power, 7-18 setting the sweep type, 5-12 setting up bandwidth test limits,
analyzer’s internal memory, 4-36 saving measurement results
ASCII data formats, 4-41 instrument state files, 4-47 scale and offset, 7-8 scale, choosing, 4-17 searching for specific amplitude,
1-92–1-94 shipment for service, 8-2 sidelobes, 3-27 single page plots, outputting
calculating, 1-42 stepped edit list menu, 7-16 stepped edit subsweep menu, 7-16 stepped list frequency sweep, 7-15
using a printer, 4-26 single point limits, 1-77 sloping limit line, 1-75 small signal transient response,
segment menu, 7-16 stepped edit list menu, 7-16 stepped edit subsweep menu,
1-39 segment menu, 7-16 segment power, setting, 7-18 selecting
7-16 stepped list mode, 1-67 stimulus state, error-correction,
measuring, 3-19 smith chart format, 7-26 smith chart markers, 1-33 smoothing, 7-8, 7-35 softkey
auto-feed, 4-15 gate shape, 3-36 line types, 4-17 pen numbers and colors, 4-16 selecting sweep modes, 7-19 sending the exit HPGL mode and
6-9 stop frequencies, minimum
allowable, 3-16 stop frequency, setting, 1-35 stopping a sequence, 1-100 stopping the ripple test, 1-86 storing
4 Param Displays, 1-18 channel position, 1-17 solving problems with printing or
form feed sequence to the printer, 4-26 sending the HPGL initialization
plotting, 4-35 solving problems with saving or
exit HPGL mode and form feed
sequence to the printer, 4-26 sending the plot file to the printer,
recalling files, 4-55 using an external disk drive,
sequence, 4-26 HPGL initialization sequence,
4-55 source attenuator switch
4-25 sequence on a disk, 1-104 sweep
changing the title, 1-103 clearing from memory, 1-102 creating, 1-98 decision making menu, 1-112 editing, 1-100 generating files in a loop counter
protection, 7-13 repetitive switching of the
rate, decreasing, 5-9 retrace, 5-7 speed, increasing, 5-10 type, setting, 5-12 sweep time, 7-11
attenuator, 7-13 source match, 7-39 source match and load match,
example, 1-116 in-depth information, 1-105 jumps to itself, 1-112 limit test example, 1-118 loading from a disk, 1-104 loop counter example, 1-115 naming files, 1-103 printing, 1-105 purging from a disk, 1-104 running, 1-100 size, 1-106 stopping, 1-100 storing to disk, 1-104 sequencing
7-70 source mismatches, minimizing,
auto sweep time mode, 7-11 manual sweep time mode, 7-11 minimum sweep time, 7-11 sweep types, 7-15
2-4 source power, setting, 1-5 source step attenuator, 7-4 S-parameter menu
CW time sweep, 7-19 linear frequency sweep, 7-15 logarithmic frequency sweep,
analog in menu, 7-22 conversion menu, 7-22 S-parameters, 7-20
7-15 power sweep, 7-19 selecting sweep modes, 7-19 stepped list frequency sweep,
S-parameter menu, 7-22 understanding, 7-20 S-parameters menu
input ports menu, 7-23 specific amplitude, 1-39
7-15 swept list frequency sweep, 7-17 sweep-to-sweep averaging, 7-6 swept edit list menu, 7-17 swept edit subsweep menu, 7-17 swept list frequency sweep, 7-17
bandwidth, searching for, 1-41 maximum amplitude, searching
special functions menu, 1-112 test, 1-98 service menu (ECal), 6-67 servicing, 8-6 setting
for, 1-39 minimum amplitude, searching
for, 1-39 target amplitude, searching for,
setting segment IF bandwidth,
7-19 setting segment power, 7-18
auto sweep time mode, 5-12
4-26 sequence · Index-10

swept edit list menu, 7-17 swept edit subsweep menu, 7-17 swept list mode
limit test example sequence,
title, display, 1-11 titling the displayed
1-118 loading a sequence from a disk,
measurement, 4-32 to produce a time stamp, 4-33 trace math operation, 7-7 trace noise, reducing, 5-16 tracking, 7-41 tracking the amplitude, 1-41 tracking, amplitude and phase,
calibrate, 1-70 characteristics of the filter, 1-68 device under test, connect, 1-67 measure, 1-70 measurement parameters, 1-68 stepped list mode, 1-67 to test a device, 1-67 swept list mode, using, 5-10 swept RF/IF conversion loss, high
1-104 loop counter example sequence,
1-115 modifying a command, 1-101 naming files generated by a
sequence, 1-103 printing a sequence, 1-105 purging a sequence from a disk,
2-39 transform, 7-8 transforming CW time
1-104 running a sequence, 1-100 stopping a sequence, 1-100 storing a sequence on a disk,
dynamic range, 2-20 switch protection, source
measurements into the frequency domain, 3-22 forward transform
attenuator, 7-13 SWR format, 7-28 SWR/return loss, 2-50 synthesized source, built-in, 7-3 system
measurements, 3-22 transmission measurements in
1-104 using to test a device, 1-114 test set, built-in, 7-4 testport output power, 5-7 text file, saving measurements as
time domain low pass, 3-19 interpreting the low pass step
bandwidth, changing, 5-16 bandwidth, widening, 5-12 system controller mode, 7-79 system operation, 7-3
transmission response horizontal axis, 3-20 interpreting the low pass step
a, 4-44 thru, manual, 6-60 time delay, decreasing, 5-9 time domain bandpass mode, 3-4,
transmission response vertical axis, 3-20 measuring separate
built-in synthesized source, 7-3 built-in test set, 7-4 microprocessor, 7-4 receiver block, 7-4 required peripheral equipment,
3-12 adjusting the relative velocity
transmission paths through the test device using low pass impulse mode, 3-20 measuring small signal
factor, 3-12 reflection measurements using
bandpass mode, 3-12 transmission measurements
7-4 systematic errors, 7-41
transient response using low pass step, 3-19 transmission measurements
using bandpass mode, 3-14 time domain low pass impulse
T taking care of microwave
mode, 3-4 time domain low pass mode, 3-15
using bandpass mode, 3-14 interpreting the bandpass
connectors, 5-3 talker/listener mode, 7-79 target amplitude, searching for,
transmission response horizontal axis, 3-14 interpreting the bandpass
using low pass, 3-18 minimum allowable stop
1-40 temperature drift, 5-5 terminology, TRL, 7-68 test
frequencies, 3-16 reflection measurements in time
transmission response vertical axis, 3-14 transmission measurements,
domain low pass, 3-16 setting frequency range for time
bandwidth, 1-92–1-97 ripple limit, 1-81–1-91 test port coupling, 7-10 test port input power, increasing,
response and isolation error correction, 6-17 transmission measurements,
domain low pass, 3-15 transmission measurements in
time domain low pass, 3-19 time domain low pass step mode,
response error correction, 6-14 transmission response
5-15 test sequencing, 1-98
3-4 time domain measurements,
measurements, making, 3-5 TRL calibration, performing, 6-52 TRL error correction
changing the sequence title,
introduction, 3-3 forward transform mode, 3-4 time domain bandpass mode,
1-103 clearing a sequence from
memory, 1-102 creating a sequence, 1-98 editing a sequence, 1-100 generating files in a loop counter
assigning standards to various
TRL classes, 6-51 label the calibration kit, 6-51 label the classes, 6-51 performing the TRL calibration,
3-4 time domain low pass impulse
mode, 3-4 time domain low pass step
example, 1-116 in-depth sequencing
mode, 3-4 time stamp, 4-33 title, 1-106
6-52 TRL error-correction, 6-50
information, 1-105 inserting a command, 1-101
fault location measurements · Index-11

calibration kit, 6-50 TRL options, 7-76 TRL terminology, 7-68 TRL* error model, 7-68 TRL*/LRM* calibration, 7-67
creating a user-defined TRL
U uncoupling display markers, 1-31 understanding power ranges, 7-9 understanding S-parameters,
7-20 upper stopband parameters, 1-69 using continuous correction mode,
calibration standards for TRL/LRM, 7-74 how TRL*/LRM* works, 7-68 improving raw source match and
6-38 using direct sampler access
configurations, 5-17 using external calibration, 5-13 using fast 2-port calibration, 5-13 using sample-and-sweep
load match for TRL*/LRM* calibration, 7-71 isolation, 7-69 source match and load match,
correction mode, 6-36 using swept list mode, 5-10
detecting IF delay, 5-11 using the parallel port, 7-80
7-70 TRL calibration procedure, 7-72 TRL options, 7-76 TRL standards, requirements,
copy mode, 7-80 GPIO mode, 7-81
7-72 TRL terminology, 7-68 TRL* error model, 7-68 TRL*/LRM* two-port calibration,
V vector error-correction, 7-7 verification, performance, 5-5 verifying performance, 7-65 vertical axis, 3-13, 3-14, 3-16,
7-55 TRM error correction, 6-54
assigning standards to various
TRM classes, 6-55 creating a user-defined TRM
3-20, 3-22 viewing a single measurement
calibration kit, 6-54 labeling the calibration kit, 6-55 labeling the classes, 6-55 modifying the standard
channel, 5-13 viewing plot files on a PC, 4-22
using AmiPro, 4-23 using Freelance, 4-23
definitions, 6-54 performing the TRM
W what you can save to a computer,
calibration, 6-56 TTL
4-37 what you can save to a floppy disk,
I/O menu, 1-108 input decision making, 1-108 out menu, 1-112 output for controlling
4-37 what you can save to the
analyzer’s internal memory, 4-36 widening the system bandwidth,
peripherals, 1-108 tuned receiver mode, 2-27, 2-29,
7-84 in-depth description, 7-85 test setup, typical, 7-84 two sources
5-12 windowing, 3-27
finite impulse width (or rise
time), 3-27 sidelobes, 3-27
addressing and configuring,
2-30 two-port calibration, full, 7-55 two-port calibration, TRL*/LRM*,
7-55 two-port error model, 7-47 type-N connector sex
clarifying, 6-4 types of error correction, 6-10
fabricating and defining · Index-12
Pages referenced in this manual — manufacturer, support and downloads.
Owners and our team usually reply within a day — and your question helps the next person too.
Be the first to ask — other owners or our team will answer.