Page 1 of the Agilent Technologies 1141A user manual — Agilent Technologies 1141A Differential Probe and 1142A Probe Control and Power Module
Page 1 · click to enlarge
Page 1

User and Service Guide

Publication number 01141-97002 July 2004

For Safety and Regulatory information, see the pages at the end of this book.

© Copyright Agilent Technologies 2000-2004

Agilent Technologies 1141A Differential Probe and 1142A Probe Control and Power Module

All Rights Reserved

Page 2 of the Agilent Technologies 1141A user manual — Agilent Technologies 1141A Differential Probe and 1142A Probe Control and Power Module
Page 2 · click to enlarge
Page 2

Agilent Technologies 1141A Differential Probe and 1142A Probe Control and Power Module

This manual contains information for use and service of the differential probe system, the 1141A Differential Probe and 1142A Probe Control and Power Module. In this document, the two models will be treated as a system.

Each of the two instrument models that make up the differential probe system has a serial number sticker. The sticker for the 1141A Differential Probe is inside the probe, in the bottom cover. (See chapter 3 for disassembly procedure.)

The 1141A/1142A probe system allows measurement of small differential signals in the presence of much larger common-mode signals. It has the following major features:

  • 200 MHz bandwidth
  • Variable offset
  • dc reject
  • ac coupling
  • Remote operating capability

The variable offset capability can be used to measure small ac signals in the presence of much larger dc levels. Remote operation of key features allows the use of the probe system in automatic test situations.

The probe can be used with an oscilloscope, spectrum analyzer, or any instrument where differential probing is required and a compatible 50 Ω input is available.

Page 5 of the Agilent Technologies 1141A user manual — Operating the Probe
Page 5 · click to enlarge
Page 5

Operating the Probe

Page 6 of the Agilent Technologies 1141A user manual — Accessories Supplied
Page 6 · click to enlarge
Page 6

Operating the Probe Accessories Supplied

Introduction

This chapter shows you how to connect and operate the 1141A Differential Probe and 1142A Probe Control and Power Module as a differential probe system.

Accessories Supplied

The following items are supplied as part of the 1141A/1142A probe system. Item numbers refer to the numbers in Figure 1-1 on page -7 and Figure 1-2 on page -8. Those without item numbers are supplied but not shown in figures. See the Replaceable Parts List for parts not listed below.

ItemDescriptionQtyPart Number 1Differential Probe1 210x Attenuator Adapter15063-2144 3100x Attenuator15063-2145 4ac Coupling Adapter15063-2146 5Two-inch Extension Leads (package 5)15959-9334 6Mini Grabbers21400-1422 7Five-inch Ground Lead15061-6162 8Shielded Signal Lead101141-68702 9Test Board101141-66504 10Flat-blade Alignment Tool18710-1961 11Circuit Connection Posts (strip of 20)11251-5943 Probe Control and Power Module11142A Power Cord1see parts list Carrying case User and Service Manual1
Page 7 of the Agilent Technologies 1141A user manual — Introduction
Page 7 · click to enlarge
Page 7

Operating the Probe Accessories Supplied

Agilent Technologies 1141A — Introduction diagram (page 7)
Fig. 1 · Introduction

1141A Differential Probe and Accessories

Page 8 of the Agilent Technologies 1141A user manual — Accessories Available
Page 8 · click to enlarge
Page 8

Operating the Probe Accessories Available

Agilent Technologies 1141A — Accessories Available diagram (page 8)
Fig. 2 · Accessories Available
Agilent Technologies 1141A — Accessories Available diagram (page 8)
Fig. 3 · Accessories Available

1142A Probe Control and Power Module

Accessories Available

The following accessories can be ordered.

  • 5959-9335 Long Extension Lead (5.5 inch/14 cm), package of 5
  • 5090-4833 Mini grabber for SMT, package of 20

1141A Miscellaneous Accessories

Page 9 of the Agilent Technologies 1141A user manual — Recommended Test Equipment
Page 9 · click to enlarge
Page 9

Operating the Probe To inspect the probe

To inspect the probe

Inspect the shipping container for damage. If the shipping container or cushioning material is damaged, it should be kept until the contents of the shipment have been checked for completeness and the instrument had been checked mechanically and electrically. Accessories supplied with the instrument are listed in see “Accessories Supplied” on page 6 of this manual. If the contents are incomplete, if there is mechanical damage or defect, or if the instrument does not pass calibration tests, notify the nearest Agilent Technologies office. If the shipping container is damaged, or the cushioning materials show sign of stress, notify the carrier as well as the nearest Agilent Technologies office. Keep the shipping materials for the carrier’s inspection. The office will arrange for repair or replacement at Agilent Technologies’ option without waiting for a claim settlement.

Using the probe with other instruments

The 1141A/1142A probe system can be used with other instruments as well as oscilloscopes. You can use it with a spectrum analyzer or frequency counter, or any instrument with an input that can be terminated with 50 Ω. If you are going to use the probe system with an instrument other than an oscilloscope, you may need to set up the probe with an oscilloscope first. This will allow you to select coupling and reject modes, and set offset, so the output of the probe is compatible with signal requirements of the other instrument.

Recommended Test Equipment

The following table is a list of the test equipment required to test calibration, make adjustments, and troubleshoot this instrument. The table indicates the critical specifications of the test equipment and for which procedure the equipment is necessary. Equipment other than the recommended model may be used if it satisfies the critical specifications listed in the table.

Recommended Agilent Model/Part Use

Equipment Required Critical Specifications

Signal Generator 1 to 200 MHz, 400 m Vrms 8648A P

Function Generator 2.5 to 500 kHz, 600 m Vp-p to 16 Vp-p 33120A A

Oscilloscope >400 MHz bandwidth, 1 mV/div sensitivity 54830B P,A

dc source ± 100 mV to 7 Vdc E3632A P

Recommended Test Equipment

Page 10 of the Agilent Technologies 1141A user manual — A
Page 10 · click to enlarge
Page 10

Operating the Probe Recommended Test Equipment

Recommended Agilent Model/Part Use

Equipment Required Critical Specifications

Test Board No substitute (supplied accessory) 01141-66504 A

Load BNC Feedthrough, 50 Ω Pasternack Enterprises PE6008- 50 or Huber+Suhner 22543742

P

Cables (2) BNC, 50 Ω 36-inch 10503A P,A

Cable BNC, 50 Ω 9-inch 10502A A

Cable Type N (m) 24-inch 11500B P

Adapter Type N (m) to BNC (f) 1250-0780 A

Adapter Type N (f) to BNC (m) 1250-0077 P

Adapters (2) BNC (f) to dual banana (m) 1251-2277 P

Alignment tool Small flat blade (supplied accessory) 8710-1961/ Sprague-Goodman part number GTT-5G

A

Setting up the probe

The following paragraphs cover system preparation and initial adjustments.

Power Requirements The 1141A/1142A probe system (specifically the 1142A) requires a power source of either 90 to 132/198 to 264 Vac, 47 to 440 Hz, 25 VA maximum.

Line Voltage Selection Before applying power, verify the setting of the LINE SELECT switch on the rear panel of the 1142A. The slide switch can be set to either 115 or 230 V.

P=Calibration Tests, A=Adjustments

Page 11 of the Agilent Technologies 1141A user manual — Initial Adjustment
Page 11 · click to enlarge
Page 11
Agilent Technologies 1141A — Initial Adjustment diagram (page 11)
Fig. 4 · Initial Adjustment

WAR N ING Before connecting this instrument, the protective earth terminal of the instrument must be connected to the protective conductor of the (Mains) power cord. The Mains plug must be inserted in a socket outlet provided with a protective each contact. The protective action must be negated by the use on an extension cord (power cable) without a protective conductor (grounding). Grounding one conductor of a two- conductor outlet does not provide an instrument ground.

This instrument is provided with a three-wire power cable. When connected to an appropriate ac power outlet, this cable grounds the instrument cabinet. The type of power cable plug shipped with the instrument depends on the country of destination. The 1142A Power Control and Power Module does not have a power switch. A power switch is not required because of the low mains power requirement.

Agilent Technologies 1141A — Initial Adjustment diagram (page 11)
Fig. 5 · Initial Adjustment

Procedure 1 Use the power cord to connect the 1142A to the ac mains. 2 Connect the 1141A probe cable power connector to the PROBE connector on the rear panel of the 1142A power module. 3 Connect the output of the probe to the input of the oscilloscope. 4 Set the input impedance of the oscilloscope to 50 Ω. If the oscilloscope does not have a selectable 50 Ω input impedance, connect a 50 Ω BNC feedthrough termination between the probe output and the input of the oscilloscope. 5 If making an initial equipment setup, continue with the initial adjustment in the following section.

Initial Adjustment

For a given combination of 1141A Differential Probe and 1142A Probe Control and Power Module, you may want to adjust the Offset Null and DC Reject Gain. Typically, you need to make these adjustments only once, before the probe is first used. You can make them any time to optimize the system. These adjustments do not affect the specifications of the probe system.

1142A Rear Panel

Page 12 of the Agilent Technologies 1141A user manual — Initial Adjustment
Page 12 · click to enlarge
Page 12

Operating the Probe Recommended Test Equipment

  • Offset null zeroes the dc level at the output of the probe. The range of adjustment is about ±4 mV.
  • DC Reject Gain adjusts the gain of the dc reject circuit to accurately null the dc component of an input signal. The range of adjustment is about ±0.5%.
Agilent Technologies 1141A — Initial Adjustment diagram (page 12)
Fig. 6 · Initial Adjustment
  • Equipment Needed The following equipment is necessary for initial adjustment.
  • 5 V power supply
  • DVM that can measure 25 µV
  • 50 Ω BNC feedthrough terminator
  • 01141-66504 test board

Equipment Setup Use the following procedure to setup the differential probe system for initial adjustment.

1 Use the probe setup procedure to set up the probe system. 2 Connect a 50 Ω BNC feedthrough terminator to the output of the probe. 3 Disconnect all accessories from the input of the probe. 4 Connect the DVM to measure the dc output of the probe at the 50 Ω load.

1142A Front Panel

Page 13 of the Agilent Technologies 1141A user manual — Using the Accessories
Page 13 · click to enlarge
Page 13

Operating the Probe Recommended Test Equipment

5 Set up the 1142A: a Set the Local/Remote push button to Local. b Under DC Couple, press Zero offset. 6 Set the power supply output to 5 V. 7 Arrange a connection between the power supply and the test board. The negative terminal of the supply should connect to the shield of the test board BNC. If your power supply has standard binding posts, you can connect a banana-to-BNC adapter to the supply and connect a BNC cable between the supply and the test board.

Adjustment Warm up the 1141A for 30 minutes before making adjustments. 1 With the 1141A probe inputs unconnected, adjust Offset Null on the 1142A for a minimum reading on the DVM. The voltage swing of the adjustment is approximately ±4 mV. 2 On the 1142A, under DC Reject, press 5.0 Hz. 3 Read and record the reading on the DVM, mV. 4 Connect the probe to the test board in the position shown below (signal to + input).

Signal to + input

5 After the DVM reading stabilizes, adjust DC Reject Gain to the reading recorded in step 3. With a 5 V supply, the voltage swing is approximately ±12.5 mV. With a lower supply, the voltage swings proportionally less.

Using the Accessories

The 1141A Differential Probe and accessories are designed to provide a variety of ways to connect to circuitry and make measurements. In the descriptions, any method used to connect to the probe signal inputs also applied to the adapters. The figure below shows, in a general way, the use of accessories.

Page 14 of the Agilent Technologies 1141A user manual — Using the Accessories
Page 14 · click to enlarge
Page 14

Operating the Probe Recommended Test Equipment

Agilent Technologies 1141A — Using the Accessories diagram (page 14)
Fig. 7 · Using the Accessories
Agilent Technologies 1141A — Using the Accessories diagram (page 14)
Fig. 8 · Using the Accessories

Probe Tips Probe tips fit into the receptacles in the probe and are held in place with probe tip caps. If necessary, you can solder the probe tips into a circuit or wires can be soldered to the tips. If you solder to the probe tips, be careful not to melt the plastic probe tip caps.

No te Because of the close tolerances between the probe tip caps and probe tips, it will be difficult to separate the probe tips and caps once the probe tips have been soldered.

Ground Leads The circular end of the ground lead fits over the screw on the top side of the probe.

  • Extension Leads The extension leads provide a flexible connection between circuitry and the probe.
  • To provide a male connection to other circuitry, connect the extension lead over the probe tips.
  • To provide a female connection, remove the probe tip caps and probe tips and connect the extension leads to the probe.

Basic Accessory Connections

Page 15 of the Agilent Technologies 1141A user manual — Using the Accessories
Page 15 · click to enlarge
Page 15

Operating the Probe Recommended Test Equipment

No te Use extension leads and similar connection accessories carefully. Extension leads compromise the high-frequency specifications of the probe. CMRR is particularly sensitive to unbalanced input parameters.

  • To prevent pickup of stray fields when you use extension lead, either the ones supplied with the 1141A or others, dress them carefully as follows:
  • Connect the leads at right angles to the circuitry under test.
  • Keep the leads as parallel as possible before they connect to the probe.

Mini-Grabbers Mini-grabbers can be attached to the probe or adapter through the extension leads. 1 Remove the probe tip caps and tips. 2 Attach the extension leads to the probe or adapter. 3 Attach the mini-grabbers to the extension leads.

Circuit Connector Posts These 0.025-inch square posts can be used to connect either directly to the probe or to the extension leads. Solder the posts directly into your circuitry or use them to make extension leads that plug into the inputs of the probe or adapters.

Shielded Signal Leads The shielded signal leads allow connection to points in a circuit that are up to 10 inches apart. The leads are shielded to within approximately 1/2 inch of the end of the lead so they minimize pick-up due to stray fields from adjacent circuitry. 1 Connect the end with the ground connector to the probe pins and ground of the differential probe or adapter. 2 Connect the free ends of the leads to 0.025-inch square or 0.030-inch round pins in your circuitry or to the mini-grabbers.

No te Each lead has an input capacitance of approximately 15 pF. This capacitance may limit the bandwidth of your measurement (depending on the impedance of the circuit). Also, CMRR may be affected because of slight differences between the input capacitance of the two leads. CMRR is also affected by differences in impedance between the two measurements points.

Page 16 of the Agilent Technologies 1141A user manual — Using the Accessories
Page 16 · click to enlarge
Page 16

Operating the Probe Recommended Test Equipment

Adapters There are three adapters for use with the differential probe. Two adapters are attenuators, a 10x and a100x. The other is an ac adapter for blocking dc from the probe input. The adapters are installed on the probe after the probe tip caps and probe tips have been removed. The adapter fastens to the probe using a thumb wheel located on the underside of the adapter. the figure below shows a good way to hold the probe while attaching the adapter. 1 Remove the probe tip caps and probe tips, from the probe. 2 Fit the adapter over the end of the probe and rotate the thumb wheel with your finger until the adapter fits snugly. A snug fit is important because the ground is maintained through the thumb wheel screw. A loosely attached adapter compromises the mechanical and electrical integrity of the combination.

Agilent Technologies 1141A — Using the Accessories diagram (page 16)
Fig. 9 · Using the Accessories
Page 17 of the Agilent Technologies 1141A user manual — Using the Accessories
Page 17 · click to enlarge
Page 17

Operating the Probe Recommended Test Equipment

  • Adapter Combinations The figure below shows the allowed adapter and probe connections. There are two specific combinations that should not be used.
  • Do not attach the ac adapter between an attenuator adapter and the probe. An attenuator adapter must be terminated by the input resistance of the probe. The ac adapter isolates the probe input resistance.
  • Do not cascade two attenuator adapters. The attenuator adapters are designed to be terminated by the 1 M Ω resistance of the probe. The input resistance of the attenuator adapter is 9 M Ω for the 10x adapter and 10 M Ω for the 100x adapter.
Agilent Technologies 1141A — Using the Accessories diagram (page 17)
Fig. 10 · Using the Accessories
Agilent Technologies 1141A — Using the Accessories diagram (page 17)
Fig. 11 · Using the Accessories

Allowed Adapter Connections

Page 18 of the Agilent Technologies 1141A user manual — Coupling Functions
Page 18 · click to enlarge
Page 18

Operating the Probe Recommended Test Equipment

  • Connector Compatibility The following are general connector characteristics for the probe, adapters, and accessories.
  • The female connectors on the probe, adapters, and other accessories are designed to mate with 0.030-inch round or 0.0250-inch square pins.
  • The probe, adapter, and extension lead pins are 0.030-inch round.
  • The strip of circuit connection posts provided as an accessory has 0.025-inch square pins.
  • The mini-grabber has a 0.25-inch square pin.
  • The ground connection at the end of the probe and adapters (where the adapters fasten) accepts an M3 metric screw.

Test Board The primary use of the test board is to apply test and calibration signals to the input of the probe or adapters. Specific use of the test board is covered wherever it applies.

Grounding

Grounding is very important when probing circuitry. Improper grounding can increase the common mode signal level. This reduces the effectiveness of the differential probe. The mechanical connections at the input of the probe are ground for probe signals. The screw where the ground lead attaches (see figure 1-7) fastens to this ground. Also, the attenuator and ac adapter fasten to this ground through the screw connection and the ground is carried through each adapter to its front.

Agilent Technologies 1141A — Grounding diagram (page 18)
Fig. 12 · Grounding

Coupling Functions

There are three methods for blocking or compensating for the dc component of a signal. Each has specific advantages.

Probe Grounds

Page 19 of the Agilent Technologies 1141A user manual — Grounding
Page 19 · click to enlarge
Page 19

Operating the Probe Recommended Test Equipment

Probe System Coupling Functions

dc offset dc Reject ac Coupling Adapter

Probe alone Probe with 10x Probe with 100x

± 20 V ± 200 V ± 500 V

± 20 V ± 200 V ± 500 V

± 20 V ± 500 V ± 500 V

Set-up needed Adjust offset to put signal on screen

Select DC Reject low-frequency corner

Attach ac Coupling adapter to differential probe

Remote Control? Yes1 Yes1 No

1 Isolated external dc reference and control signals are needed

  • dc Reject dc Reject is the best method of eliminating the dc component of a signal when dc is not a factor in the measurement. The key characteristics are:
  • The low-frequency component (from dc to the selected corner frequency) is automatically nulled by the dc reject circuitry.
  • Probe CMRR specifications are not compromised as happens when the ac coupling adapter is uses.
  • There is a selectable low frequency corner with -3dB points at 0.05 Hz, 0.5 Hz, or 5.0 Hz.
  • The voltage reject range is ±20 V with the probe alone, ±200 V with the 10x attenuator, and ±500 V with the 100x attenuator.
Show 5 more labels / short text on this page
dc Blocked · CMRR degradation? · No No Yes · Low-frequency degradation? · No Yes Yes
Page 20 of the Agilent Technologies 1141A user manual — Grounding
Page 20 · click to enlarge
Page 20

Operating the Probe Recommended Test Equipment

To use dc reject: 1 Remove the ac adapter if it is installed. 2 On the front panel of the 1142A, press Local. 3 Under DC Reject on the front panel, press 5.0 Hz or 0.5 Hz individually, or 5.0 Hz and 0.5 Hz simultaneously to get 0.05 Hz. Within the frequency and voltage characteristics noted elsewhere in this manual, low frequencies are nulled from the input signal.

  • Offset Offset is the best method to use when the low-frequency corners associated with dc reject and the ac adapter interfere with the measurement. The key characteristics are:
  • The user manually null the dc component with the offset adjustment.
  • Offset is dc coupled so there is no low frequency roll-off.
  • Probe CMRR specifications are not compromised as happens when the ac coupling adapter is used.
  • The voltage offset range is ±20 V with the probe alone, ±200 V with the 10x attenuator, and ±500 V with the 100x attenuator. (With the 100x attenuator, the offset range is restricted by the maximum input voltage rating rather than the operating range of the offset). To use offset: 1 Remove the ac adapter if it is installed. 2 On the front panel of the 1142A, press Local and Variable offset. 3 Adjust the Coarse and Fine Variable Offset until the signal is displayed on the screen of the oscilloscope.
  • ac Adapter The ac adapter must be used when the dc component of the signal exceeds the operating range of the dc reject or offset methods. The ac adapter block the dc and low frequency component of the input by forming a high pass filter with the input impedance of the probe or adapter. The key characteristics are:
  • The ac adapter safely blocks ±200 Vdc when attached directly to the probe or ±500 Vdc when attached to a 10x or 100x adapter.
  • The probe and adapters have different input impedances, so they have different low frequency corners with the ac adapter. When the ac adapter is directly on the probe the -3dB corner is 15 Hz. When the ac adapter is on an attenuator the corner is 1.5 Hz.
  • The low-frequency CMRR when using the ac adapter is not as good as when using the probe alone or the probe with a 10x or 100x adapter.
Page 21 of the Agilent Technologies 1141A user manual — Remote operation
Page 21 · click to enlarge
Page 21

Operating the Probe Recommended Test Equipment

To use ac coupling: 1 Attach the ac coupling adapter to the input of the probe or the input of the attenuator adapter. 2 On the 1142A, press Local and Zero offset.

Remote operation

For automatic test applications, the coupling and offset functions provided by the 1142A Probe Control and Power Module can be remotely controlled through a connector on the rear panel of the module. The connection is through a standard 9-pin female D-subminiature connector. This style is the same as that used on some personal computer monitor cables, which provides an economical way to connect the 1142A to the controller interface on an automatic test system. The following table gives the connections.

Agilent Technologies 1141A — Remote operation diagram (page 21)
Fig. 13 · Remote operation

1 Function Select 1 (A1R) 6 Function Select 0 (A0R)

NOT E To minimize dc offset errors and potential noise coupling, electrically isolate all connections between the Remote Input connector and the controlling system.

Page 22 of the Agilent Technologies 1141A user manual — Cleaning Requirements
Page 22 · click to enlarge
Page 22

Operating the Probe Recommended Test Equipment

Function Select The easiest way to control the function select lines is contact closures between the lines and Digital common (pin 2) of the remote input connector. (TTL compatible control signals can be used; but to avoid problems with ground loops, they must be electrically isolated.) The following truth table shows the functions provided by the function select lines. For the Remote Inputs, “0” represents a closure and “1” represents an open circuit.

Remote Functions Select Truth Table

0 0 0.05 Hz Reject

0 1 0.5 Hz Reject

1 0 5 Hz Reject

1 1 DC Couple

Variable Offset The remote variable offset can be used when the dc couple function is remotely selected. The offset voltage must be referenced to the External offset common (pin 4) of the remote input connector. It must be electrically isolated from the controlling system. The following table shows the offset range and remote offset requirements for probe and adapter combinations.

Remote Offset Input Requirements

Offset Range Remote Requirements

Probe alone ± 20 V ± 10 V

Probe with 10x adapter ± 200 V ± 10 V

Probe with 100x adapter ± 500 V ± 2.5 V

Cleaning Requirements

If the instrument requires cleaning: (1) Remove power from the instrument. (2) Clean the external surfaces of the instrument with a soft cloth dampened with a mixture of mild detergent and water. (3) Make sure that the instrument is completely dry before reconnecting it to a power source.

Remote Input Function · A1R (Pin 1) · A0R (Pin 6)

Page 23 of the Agilent Technologies 1141A user manual — Differential Amplifiers and CMRR
Page 23 · click to enlarge
Page 23

Operating the Probe Recommended Test Equipment

Differential Amplifiers and CMRR

  • The 1141A Differential Probe is a high-impedance differential amplifier. A characteristic of differential amplifiers is the ability to reject signals that are common to the two inputs. The common mode rejection ratio (CMRR) is the measurement of this ability. It is expressed as the ratio between the amplitudes of the common mode and differential signals which product equal outputs. For example, if a common mode signal of 1 V and differential signal of 1 mV both produce outputs of 1 mV, the CMRR is 1000:1. The ability to reject common mode signals is dependent on the balance designed into the differential amplifier. At higher frequencies it becomes harder to balance circuit parasitics and parameters of devices so CMRR degrades as frequency increases. Also, stray coupling increases with frequency and coupling may vary between the two differential paths. The CMRR of the 1141A Differential Probe is specified at the input of the probe and cannot be affected expect by adjustments in the probe. However, the way the probe is connected into the circuitry being tested can have a big influence in the overall result of the measurement, especially at high frequencies. The following things can affect the effective CMRR of a test setup:
  • The connection to the circuit under test. The method used to connect the probe is important because it involves the symmetry of the differential input circuitry. For example, using different lengths of wire to connect the circuit to the two probe inputs unbalances the inductance and capacitance at the inputs. The effective CMRR will be reduced, especially at high frequencies. Additionally, coupling from adjacent circuitry will be less balanced.
  • The impedance of the source. This is another instance where the symmetry of the differential circuit is important. The impedance of the source forms a network with the input impedance of the source forms a network with the input impedance of the connections and the probe. This network determines the frequency response for the measurement. If each side of the differential source has a different impedance, the frequency response of each side will be different and the unbalance is reflected in a reduced CMRR. Of course, lower source impedances have less effect on the frequency response of the measurement.
  • The ground connection. A poorly located ground connection allows ground loops to add to the common mode signal.
  • Frequency. Frequency is the most important factor in CMRR only because all of the factors mentioned above are frequency dependent. The unbalances of capacitance and inductance are more important as frequency increases. Therefore, good highfrequency practice is important when using a high impedance differential probe. On the other hand, if the differential probe is ac coupled to the circuit under test (the ac adapter is being used) the CMRR will be degraded below a certain frequency; the lower the frequency the worse the CMRR. This is because unbalance in the series capacitances of the ac coupler becomes more significant the lower the frequency.
Page 24 of the Agilent Technologies 1141A user manual — Differential Amplifiers and CMRR
Page 24 · click to enlarge
Page 24

Operating the Probe Recommended Test Equipment

Page 25 of the Agilent Technologies 1141A user manual — Calibration Tests and Adjustment
Page 25 · click to enlarge
Page 25

Calibration Tests and Adjustment

Page 26 of the Agilent Technologies 1141A user manual — The Test Board
Page 26 · click to enlarge
Page 26

Calibration Tests and Adjustment Equipment Required

Introduction

This chapter is divided into two sections. The first section gives calibration tests and the second adjustment procedures for the 1141A Differential Probe and 1142A Probe Control and Power Module.

Equipment Required

A complete list of equipment required for the calibration tests and adjustments is listed in “Recommended Test Equipment” on page 9. Equipment required for individual procedures is listed at the procedure. Any equipment satisfying the critical specifications listed may be substituted for the recommended model.

The Test Board

The test board is a supplied accessory for use during calibration tests and adjustments to connect signals to the differential probe (with or without adapters). A BNC connector connects the test board to a cable from the signal generator. The board includes a 50 Ω termination (two 100 Ω resistors).

Once the probe tip caps and probe tips have been removed, the probe can be connected to the test board in one of three ways, as shown in the figure below. At each position of the probe, probe inputs are connected to a different combination of signal and ground. A separate terminal on the test board connects the ground of the probe to the signal ground.

Test Board Showing Probe Positions

Page 27 of the Agilent Technologies 1141A user manual — The Test Board
Page 27 · click to enlarge
Page 27

Calibration Tests and Adjustment

The Test Board

Calibration Tests

These procedures test the probe’s electrical performance using applicable specifications given in “Performance Specifications and Characteristics” on page 55 as performance standards. Specifications applicable to individual tests are noted at the test for reference.

Testing Interval The calibration testing procedures may be performed for incoming inspection of the instrument and should be performed periodically thereafter to ensure and maintain peak performance. The recommended test interval is yearly or every 2,000 hours of operation. Amount of use, environmental conditions, and the user’s experience concerning need for testing will contribute to verification requirements.

Calibration Test Record The results of the calibration tests may be tabulated in the Calibration Test Record provided at the end of the calibration tests. The Calibration Test Record listed the calibration tests and provides an area to mark test results. The results recorded in the Calibration Test Record during initial inspection may be used for later comparisons of the tests during periodic maintenance, troubleshooting, and after repairs or adjustments.

Calibration Test Procedures Procedures may be done individually or in any order.

NOT E Allow the instrument to warm up for at least 30 minutes prior to beginning calibration tests.

Page 28 of the Agilent Technologies 1141A user manual — dc Gain Accuracy
Page 28 · click to enlarge
Page 28

Calibration Tests and Adjustment dc Gain Accuracy

dc Gain Accuracy

This test checks the dc gain accuracy of the differential probe and the dc accuracy of the differential probe with attenuator and adapters.

SpecificationProbe alone, ±2%; with attenuator adapter, ±4%

Equipment Required Critical Specifications

ac/dc Calibrator or dc Power Supply

100 mV to 7 V

100 mV to 7 V E3632A

DVM 0.5% accuracy, 10 µV resolution 3458A or E34401A

Load BNC Feedthrough, 50 Ω Pasternack Enterprises PE6008-50 or Huber+Suhner 22543742

Cables (2) BNC 50 Ω 10503A

Adapters (2) BNC (f) to dual banana (m) 1251-2277

Test Board No substitute 01141-66514

Probe Gain Test Procedure 1 Connect the probe and test equipment as shown in figure 2-2.

Equipment Required · Recommended Agilent Model/Part

Page 29 of the Agilent Technologies 1141A user manual — dc Gain Accuracy
Page 29 · click to enlarge
Page 29

Calibration Tests and Adjustment

dc Gain Accuracy

Agilent Technologies 1141A — dc Gain Accuracy diagram (page 29)
Fig. 14 · dc Gain Accuracy

2 Set up the 1142A probe control and power module as follows: a Set the Local/Remote push button to Local. b Under DC Couple, press the Zero offset button. 3 With the 1141A Probe Amp disconnected from the test PCA, adjust the Offset Null control on the 1142A until the DVM reads 0Vdc. If the probe output voltage cannot be set to 0V, subtract this voltage from the subsequent measurements in this test. 4 Connect the input of the Probe Amp to the test board in the position shown in Figure 2-4 on page -31. 5 Adjust the DC Source to output 100 mV (nom.) 6 Record the Vin1 measurement from the top DVM in figure 2-2. 7 Record the Vout1 measurement from the bottom DVM in figure 2-2. 8 Connect the Probe Amp to measure a negative voltage as shown in figure 2-3. 9 Record the Vout2 measurement from the bottom DVM in figure 2-2

Page 30 of the Agilent Technologies 1141A user manual — 2 V in 1 × --------------------------------- =
Page 30 · click to enlarge
Page 30
Agilent Technologies 1141A — dc Gain Accuracy diagram (page 30)
Fig. 15 · dc Gain Accuracy

Signal to - input

10 Calculate probe gain as

∆ V out

∆Vin Vout1 Vout2 – ( )

2 V in 1 × =

Record the result of this calculation in the “Calibration Test Record” on page 39. To pass this test, the probe gain = 0.98 to 1.02

NOT E Failure of the gain accuracy test can be caused by mis-adjustment of the probe. Perform the Probe Adjustment procedure in the Adjustments section later in this chapter and retest.

10x Attenuator Accuracy Test

NOT E If the gain test for the probe fails, the 10x Attenuator Accuracy Test will fail or the results will be poor. Do not continue until the probe passes the dc gain test.

1 Disconnect the probe from the test board and connect the 10x attenuator adapter to the probe. 2 Carefully connect the input of the probe/attenuator to the test board in the position shown in the figure below (signal to + input).

Page 31 of the Agilent Technologies 1141A user manual — 2 V in 1 × --------------------------------- =
Page 31 · click to enlarge
Page 31
Agilent Technologies 1141A — dc Gain Accuracy diagram (page 31)
Fig. 16 · dc Gain Accuracy

Signal to + input

3 Set the dc calibrator output 3 V dc. 4 Record the Vin1 measurement from the top DVM in figure 2-2. 5 Record the Vout1 measurement from the bottom DVM in figure 2-2 6 Carefully connect the input of the probe/attenuator to the test board in the position shown in the figure below (signal to - input).

Signal to - input

7 Record the Vout2 measurement from the bottom DVM in figure 2-2. 8 Calculate the 10x attenuator gain as

∆ V out

∆Vin Vout1 Vout2 – ( )

2 V in 1 × =

Record the result of this calculation in the “Calibration Test Record” on page 39.

NOT E Failure of the accuracy test for the 10x attenuator can be caused by mis-adjustment of the low-frequency CMRR (LF CMRR) adjustment. Perform the Attenuator Adapter Adjustment procedure in the Adjustments section later in this chapter then retest the attenuator adapter. If if continues to fail, repair is necessary.

Page 32 of the Agilent Technologies 1141A user manual — dc Gain Accuracy
Page 32 · click to enlarge
Page 32

Calibration Tests and Adjustment dc Gain Accuracy

100x Attenuator Accuracy Test

NOT E If the gain test for the probe fails, it will be reflected in the test for the 100x attenuator adapter. Do not continue until the probe passes the gain test.

1 Disconnect the probe/attenuator from the test board. Remove the 10x attenuator adapter from the probe and connect the 100x attenuator adapter. 2 Carefully connect the input of the probe/attenuator to the test board in the position shown in the figure below (signal to + input).

Signal to + input

3 Set the dc source output to 7 Vdc. 4 Record the Vin1 measurement from the top DVM in figure 2-2. 5 Record the Vout1 measurement from the bottom DVM in figure 2-2

Page 33 of the Agilent Technologies 1141A user manual — ∆ V out
Page 33 · click to enlarge
Page 33

Calibration Tests and Adjustment

dc Gain Accuracy

6 Carefully connect the input of the probe/attenuator to the test board in the position shown in the figure below (signal to - input).

Signal to - input

7 Record the Vout2 measurement from the bottom DVM in figure 2-2. 8 Calculate the 100x attenuator gain as

∆Vin Vout1 Vout2 – ( )

∆ V out

2 V in 1 × --------------------------------- =

Record the result of this calculation in the “Calibration Test Record” on page 39

NOT E Failure of the accuracy test for the 100x attenuator can be caused by mis-adjustment of the low-frequency CMRR (LF CMRR) adjustment. Perform the Attenuator Adapter Adjustment procedure in the Adjustments section later in this chapter then retest the attenuator adapter. If if continues to fail, repair is necessary.

Page 34 of the Agilent Technologies 1141A user manual — Bandwidth
Page 34 · click to enlarge
Page 34

Calibration Tests and Adjustment Bandwidth

Bandwidth

This test checks the high-frequency response of the 1141A Differential Probe. The bandwidth of the oscilloscope is characterized first so it is not a factor in the measurement.

Specification (-3dB, dc coupled)dc to 200 MHz

Critical Specifications Recommended Model/Part

Oscilloscope 400 MHz bandwidth 54830A

Signal Generator 200 MHz at ≈ 230 mVrms 8648A

Test Board No substitute 01141-66504

Cable Type N (m) 24-inch 11500B

Adapter Type N (f) to BNC (m) 1250-0077

Procedure This test depends on the accuracy of the termination on the test board and the termination in the oscilloscope. Both should be with 1%. 1 With the N cable and N-to-BNC adapter, connect the signal generator to the oscilloscope channel 1 input. 2 Set the signal generator for 200 MHz at 0 dBm (about 224 mVrms). 3 Set the 1142A front panel switches to Local and Zero offset. 4 On the oscilloscope press AUTOSCALE, then set the following parameters.

TIMEBASE (time/div) 2 ns/div

CHAN 1 (sensitivity) (input R)

100 mV/div 50 Ω DC

ACQUISITION Sampling Mode Memory Depth Sample Rate Averaging # of avg

Real Time Automatic Automatic Enabled 32

5 The signal on screen should be about six divisions amplitude. Measure the peak-to-peak voltage of the channel 1 signal and record the reading. Vp-p (1) = mV

Equipment Required · Equipment Required · Menu Selection Setting

Page 35 of the Agilent Technologies 1141A user manual — Bandwidth
Page 35 · click to enlarge
Page 35

Calibration Tests and Adjustment

Bandwidth

6 Reconfigure the equipment. a Disconnect the signal generator cable from the oscilloscope input and connect it to the test board. b Connect the output of the differential probe to the channel 1 input of the oscilloscope. c Carefully connect the input of the probe to the test board in the position shown in the figure below (signal to + input).

Signal to + input

7 Record the Vp-p reading on the oscilloscope. Vp-p (1) = mV 8 Divide the reading from step 6 by the reading from step 4.

Answer from step 6 Answer from step 4 =

Record the result in the Calibration Test Record. The result should be 0.707 or greater, indicating a probe bandwidth of 200 MHz or more. The bandwidth can be checked at other signal levels. Change the signal generator output level and oscilloscope V/div range proportionally.

NOT E Failure of the bandwidth test can be caused by mis-adjustment of the probe. Perform the Probe Adjustment procedure in the Adjustments section later in this chapter.

Page 36 of the Agilent Technologies 1141A user manual — CMRR Test
Page 36 · click to enlarge
Page 36

Calibration Tests and Adjustment CMRR Test

CMRR Test

This test checks the CMRR at 1 MHz and 100 MHz.

Specification 3000:1 at 1 MHz, 10:1 at 100 MHz

Agilent Technologies 1141A — CMRR Test diagram (page 36)
Fig. 17 · CMRR Test

Signal Generator 1-100 MHz at = 400 mVrms 8648A

Procedure 1 Connect the probe power connector to the PROBE connection on the rear of the 1142A Probe Control and Power Module. 2 Set the 1142A front panel switches to Local and Zero offset. 3 Connect the probe output to the oscilloscope channel 1 input. 4 Connect the input of the probe to the test board in the position shown in the figure below (signal to + input).

Signal to + input

5 Connect the signal generator to the test board. 6 Set the signal generator for 1 MHz at 385 mVrms (1 Vp-p).

Page 37 of the Agilent Technologies 1141A user manual — CMRR
Page 37 · click to enlarge
Page 37

Calibration Tests and Adjustment

CMRR Test

7 On the oscilloscope, press AUTOSCALE and set the following parameters.

Agilent Technologies 1141A — CMRR Test diagram (page 37)
Fig. 18 · CMRR Test

ACQUISITION Sampling Mode Memory Depth Sample Rate Averaging # of avg

8 On the oscilloscope, measure the peak-to-peak voltage of the channel 1 signal, then V P-P, then press 1) and record the reading. Vp-p (1) = mV 9 Connect the input of the probe to the test board in the position shown in the figure below (signal to both inputs).

Signal to both input

10 Set the sensitivity to 1 mV/div. 11 After the measurement settles (averaging is complete), record the V P-P reading. Vp-p (2) = mV 12 Disconnect the probe amp from the test board and measure Vnoise pp on channel 1. 13 Calculate the CMRR result as follows

Vpp1 Vpp2 Vnoisepp – =

CMRR

14 The result in step 12 should be ≥ 3000, representing a CMRR of 3000:1 or more. Record the CMRR in the Calibration Test Record.

Page 38 of the Agilent Technologies 1141A user manual — CMRR
Page 38 · click to enlarge
Page 38

Calibration Tests and Adjustment CMRR Test

15 Connect the input of the probe to the test board in the position shown in the figure below (signal to + input).

Signal to + input

16 Set the signal generator for 100 MHz at 0.0 dBm (about 224 mVrms, 632 mVp-p). 17 Set the oscilloscope to channel 1 and change the horizontal scale to 5 ns/div. 18 After the measurement settles (averaging is complete), note the V P-P reading. Vp-p (1) = mV 19 Connect the input of the probe to the test board in the position shown in the figure below (signal to both inputs).

Signal to both input

20 Set the channel 1 sensitivity to 10 mV/div. 21 After the measurement settles (averaging is complete), not the V P-P reading. Vp-p (2) = mV 22 Disconnect the probe amp from the test board and measure Vnoise pp on channel 1. 23 Calculate the CMRR result as follows

Vpp1 Vpp2 Vnoisepp – =

CMRR

24 The result in step 21 should be ≥ 10, representing a CMRR of 10:1 or more. Record the CMRR in the Calibration Test Record.

Page 39 of the Agilent Technologies 1141A user manual — Test Limits Results
Page 39 · click to enlarge
Page 39

Calibration Tests and Adjustment

Calibration Test Record

Calibration Test Record

1141A/1142A Differential Probe Tested by

Serial No Work Order No

Recommended Test Interval - 1 Year/2000 hours Date

Recommended next testing Temperature

Test Limits Results

dc Gain Accuracy Probe Only

+0.98 mV to +1.02 mV

10x +0.096 mV to +0.104 mV

100x +0.0096 mV to 0.0104 mV

Bandwidth >0.707 at 200 MHz

CMRR 1 MHz ≥3000:1

100 MHz ≥10:1

Page 40 of the Agilent Technologies 1141A user manual — Probe Adjustment
Page 40 · click to enlarge
Page 40

Calibration Tests and Adjustment Probe Adjustment

Adjustments

This section provides adjustment procedures for the 1141A Differential Probe and attenuator adapters. There are no service adjustments for the 1142A Probe Control Module.

Adjustment Interval None of the adjustment procedures that follow should be considered for a routine maintenance plan. The differential probe and attenuator adapters should be adjusted under conditions specified at the beginning of the respective procedures.

NOT E Warm up the instrument for 30 minutes before starting adjustment procedures.

Probe Adjustment

This procedure adjust the high-frequency and low-frequency paths on the 1141A Differential Probe.

NOT E Do not perform this procedure as a part of routine maintenance. Perform the procedure only if the probe does not meet specifications or has been repaired.

CAUTION You are going to remove the covers of the probe, so the assembly inside will be exposed while under power. The PC assembly will be electrically and mechanically vulnerable. Do these adjustment procedures in an ESD-safe area. Avoid inadvertent contact between the powered assembly and nearby tools and equipment. Avoid mechanical damage by carefully handling the exposed assembly and cables.

Page 41 of the Agilent Technologies 1141A user manual — Probe Adjustment
Page 41 · click to enlarge
Page 41

Calibration Tests and Adjustment

Probe Adjustment

The following equipment is required for this procedure. Procedures are based on the model or part number recommended.

Equipment Required Critical Specifications Recommended Agilent Model/Part

Function Generator 2.5 kHz, 1 Vp-p 33120A

Oscilloscope 300 MHz at 1 mV/div 54830B

Signal Generator 200 MHz at ≈ 300 mVrms 8648A

Test Board No substitute 01141-66504

Cables (2) BNC, 50 Ω 10503A

Adapter Type N (m) to BNC (f) 1250-0780

Probe Preparation The probe cover must be removed before adjustment. Drift due to temperature differences with and without covers is negligible. 1 Remove the probe tip caps and probe tips. 2 Loosen the probe clamp ring at the cable end of the probe (1/4 turn counter- clockwise) and slide it down the cable. 3 Remove the bottom cover. a At the cable end of the probe, separate the covers about centimeter (1/2 inch). b Slide the bottom cover toward the cable end of the probe until the locator pins at the probe input clear the holes. Then, remove the cover.

NOT E Note the position of the ground block at the input end of the probe. The ground block is held, through the PC assembly, by the grounding screw on the top of the probe. The ground block must be reinstalled on the PC assembly after the top cover is removed. Handle the PC assembly by the edges of the PC board.

4 Remove the ground connection screw on the top of the probe. The ground block will become free. At the cable end, the PC board fits over pins inside the top cover. 5 Lift the board off of the pins in the cover and slide it in the direction of the cable until the input connectors clear the front of the probe.

Equipment Required

Page 42 of the Agilent Technologies 1141A user manual — Adjustment Procedure
Page 42 · click to enlarge
Page 42

Calibration Tests and Adjustment Adjustment Procedure

6 As shown in the figure below, use the grounding screw to reinstall the ground block on the PC assembly.

Agilent Technologies 1141A — Adjustment Procedure diagram (page 42)
Fig. 19 · Adjustment Procedure

Attaching Ground Block to Probe PC Assembly

The ground block provides a mechanical and electrical connection when the probe PC assembly is connected to the test board. 7 Connect the probe power connector to the PROBE connection on the rear of the 1142A Probe Control and Power Module. 8 Connect the mains power to the 1142A. 9 Set the 1142A front panel switches to Local and Zero offset.

Adjustment Procedure

Unless specified elsewhere, the procedures must be followed in the order given. The only adjustment which may be done separately is HF COMP, the high-frequency compensation.

  • HF Gain and HF CMRR This adjustment sequence adjusts the HF Gain for unity gain at 500 kHz and the HF CMRR for minimum with a 500 kHz common mode signal. 1 Set up the function generator.
  • Sine wave
  • 500 kHz
  • 600 mV p-p 2 Use BNC cables to connect the function generator to the oscilloscope.
  • Generator OUTPUT to oscilloscope channel 1 input
  • Generator TRIG OUTPUT to oscilloscope EXT TRIG
Page 43 of the Agilent Technologies 1141A user manual — Adjustment Procedure
Page 43 · click to enlarge
Page 43

Calibration Tests and Adjustment

Adjustment Procedure

3 On the oscilloscope, then press CLEAR DISPLAY. Press AUTOSCALE, then set up the following parameters.

Agilent Technologies 1141A — Adjustment Procedure diagram (page 43)
Fig. 20 · Adjustment Procedure

ACQUISITION Sampling Mode Memory Depth Sample Rate Averaging # of avg

4 On the oscilloscope, measure the peak-to-peak voltage of the channel 1 signal and record the reading. Vp-p (1) = mV 5 Disconnect the BNC cable from the channel 1 input and connect it to the BNC connector on the test board. 6 Connect the output of the probe to the channel 1 input. 7 Carefully connect the input of the probe to the test board in the position shown in the figure below (signal to + input).

Signal to + input

Page 44 of the Agilent Technologies 1141A user manual — Adjustment Procedure
Page 44 · click to enlarge
Page 44

Calibration Tests and Adjustment Adjustment Procedure

8 Center adjustment R11, HF CMRR (see following figure).

Agilent Technologies 1141A — Adjustment Procedure diagram (page 44)
Fig. 21 · Adjustment Procedure

R11, HF CMRR Adjustment

9 Adjust R9, HF GAIN so the Vp-p (1) measurement is the same as in step 4, within ±0.5%. Make the adjustment slowly so the oscilloscope display has time to react to signal averaging. Press CLEAR DISPLAY occasionally to restart averaging, which gives a quicker indication of changes. 10 Carefully connect the probe to the test board in the position shown in the figure below (signal to both inputs).

Signal to both inputs

11 Set the function generator output to 1.0 Vp-p. 12 On the oscilloscope, set the channel 1 sensitivity to 1.00 mV/div. 13 Adjust R11 for minimum signal amplitude as shown in Vp-p reading for channel 1. Adjust R11 slowly and use CLEAR DISPLAY frequently to restart averaging.

  • Low Frequency Response and CMRR This adjustment sequence continues from the HF Gain and HF CMRR adjustments. Adjust R14 and C4 for pulse response, and adjust C6 for low-frequency CMRR. 1 Change the function generator settings to:
  • Square wave
  • 2.5 kHz
  • 600 mV p-p
Page 45 of the Agilent Technologies 1141A user manual — Adjustment Procedure
Page 45 · click to enlarge
Page 45

Calibration Tests and Adjustment

Adjustment Procedure

2 Change the oscilloscope settings to:

TIMEBASE (time/div) 50 µs/div

CHAN 1 (sensitivity) 100 mV/div

3 Carefully connect the input of the probe to the test board in the position shown in the figure below (signal to + input).

Signal to + input

4 Adjust R14 (LF Gain) and C4 (+ LF BANDWIDTH) for the flattest pulse top (see figure below). Again, adjust slowly and press CLEAR DISPLAY frequently to restart averaging.

Agilent Technologies 1141A — Adjustment Procedure diagram (page 45)
Fig. 22 · Adjustment Procedure

R14 and C4 Adjustment

Menu Selection Setting

Page 46 of the Agilent Technologies 1141A user manual — Adjustment Procedure
Page 46 · click to enlarge
Page 46

Calibration Tests and Adjustment Adjustment Procedure

5 Carefully connect the probe to the test board in position shown in the figure below (signal to both inputs).

Signal to both inputs

  • 6 Change the function generator settings to:
  • Sine wave
  • 4 kHz
  • 1.0 V p-p 7 Change the oscilloscope settings to:

TIMEBASE (time/div) 50 µs/div

CHAN 1 (sensitivity) 2 mV/div

8 Adjust C6 (-LF BANDWIDTH) for minimum signal amplitude on the oscilloscope. Again, adjust C6 slowly and press CLEAR DISPLAY frequently to restart averaging.

Menu Selection Setting

Page 47 of the Agilent Technologies 1141A user manual — Adjustment Procedure
Page 47 · click to enlarge
Page 47

Calibration Tests and Adjustment

Adjustment Procedure

High Frequency Compensation This adjustment sequence continues from the Low Frequency Response and CMRR adjustment. However, it can be done separately if the probe meets all specifications except bandwidth. Adjust R13 for unity gain at 200 MHz. 1 Connect the signal generator to the test board and set it for 200 MHz and 300mVp-p (107 mVrms). 2 Carefully connect the input of the probe to the test board in the position shown in the figure below (signal to + input).

Signal to + input

3 Press AUTOSCALE, then measure the peak-to-peak voltage on channel 1 (Press SHIFT (blue), press V P-P, then press 1). 4 Adjust R13 (HF COMP) to make the signal amplitude measurement on the oscilloscope 300 mVp-p, or as close to that as possible. Adjust slowly and press CLEAR DISPLAY frequently to restart averaging.

Agilent Technologies 1141A — Adjustment Procedure diagram (page 47)
Fig. 23 · Adjustment Procedure

The minimum allowable amplitude is 212 mVp-p. Typical values will be between 275 and 325 mVp-p. the probe needs repair if the minimum cannot be reached.

Probe Reassembly 1 Disconnect the probe power cable at the rear panel of the 1142A 2 Remove the probe PC assembly from the test board. Be sure the probe input connectors remain attached to the probe. 3 Remove the grounding screw and ground block from the PC assembly.

R13 Adjustment

Page 48 of the Agilent Technologies 1141A user manual — Adjustment Procedure
Page 48 · click to enlarge
Page 48

Calibration Tests and Adjustment Adjustment Procedure

4 Assemble the PC assembly in the top cover. The side of the PC assembly with the large hybrid is exposed when the assembly is in the top cover. The figure below shows how the top cover, PC board, and ground block fit together.

Agilent Technologies 1141A — Adjustment Procedure diagram (page 48)
Fig. 24 · Adjustment Procedure

a Insert the input connectors first, and seat the cable end of the PC assembly over the pins at the rear of the cover. b Position the ground block at the center-front of the PC assembly. c Insert the grounding screw through the top cover and PC assembly and screw it into the ground block as shown. 5 Replace the bottom cover. a Position the cable strain relief and with one hand, hold the cable and top together. The flange on the strain relief has a notch that fits around a protrusion in the top cover.

b Insert the pins at the front of the bottom cover into the holes at the front of the top cover. c Close the two covers together and fasten with the probe clamp ring.

Reassembling the Probe

Page 49 of the Agilent Technologies 1141A user manual — Attenuator Adapter Adjustment
Page 49 · click to enlarge
Page 49

Calibration Tests and Adjustment

Attenuator Adapter Adjustment

Attenuator Adapter Adjustment

The following procedure should be used if it is necessary to adjust an attenuator adapter. Attenuator adapters have only characteristics; they do not have any specifications. An adapter will need adjustment only if one or more of the following occurs.

  • If an adapter is to be used on a different 1141A probe that it was calibrated with last.
  • If an adapter needs to be optimized to requirements for a special measurement.
  • If an adapter is suspected of needing adjustment.

NOT E Attenuator adapters should not be adjusted as part of routine maintenance. The design necessary to give the high CMRR and high impedance of the attenuators makes adjustment delicate. Additionally, because of broadband noise, a spectrum analyzer should be used to verify adjustment quality.

Each attenuator contains two identical voltage dividers on a ceramic substrate. There are three adjustments. The low-frequency CMRR adjust the low-frequency balance between the two voltage dividers. The two high-frequency adjustments are electrically identical. Each adjusts the high-frequency compensation of one of the voltage dividers. To meet both pulse response and CMRR characteristics they are adjusted differently. Briefly, the adjustment procedure is:

a Adjust the positive high-frequency response (+ HF RESP) for the best pulse response using a 3.5 kHz square wave. b Adjust the negative high-frequency response (-HF RESP) and Low-frequency CMRR (LF CMRR) for best CMRR using a 3.5 kHz square wave. The following equipment is required for this procedure. Procedures are based on the model or part number recommended.

Equipment Required Critical Specifications Recommended Model/Part

Function Generator 3.5 kHz, 3 V to 16 Vp-p 33120A

Oscilloscope 100 MHz at 1 mV/div 54830B

Probe/Power Module No substitute 1141A/1142A

Test Board No substitute 01141-66504

Cable BNC, 50 Ω 9-inch 10502A

Cable BNC, 50 Ω 36-inch 10503A

Alignment tool Flat blade (supplied accessory) 8710-1961/ Sprague-Goodman part, GTT-5G

Equipment Required

Page 50 of the Agilent Technologies 1141A user manual — Attenuator Adapter Adjustment
Page 50 · click to enlarge
Page 50

Calibration Tests and Adjustment Attenuator Adapter Adjustment

NOT E The attenuator must be adjusted when installed on the 1141A probe with which it will be used. The specifications and characteristics will not be met if the attenuator adapter is adjusted with one differential probe then used with another.

  • 1 Remove the probe pins from the attenuator adapter and differential probe, then attach the adapter to the probe. 2 Set the 1142A front panel switches to Local and Zero offset. 3 Use the 9-inch BNC cable to connect the function generator to the test board. The short cable minimized ground-loop voltages. 4 Set up the function generator.
  • Square wave
  • 3.5 kHz
  • 3.0 Vp-p for 10x adapter and 16 V p-p for a 100x adapter. 5 Use the long BNC cable to connect the Trig Out of the function generator to the EXT TRIG of the oscilloscope. 6 Set up the oscilloscope, then set the following parameters.

TIMEBASE (time/div) 50 µs/div

CHAN 1 (sensitivity) (input R)

50 mV/div 50 Ω DC

TRIG (mode) source level

trg’d EXT 1.00000 V

ACQUISITION Sampling Mode Memory Depth Sample Rate Averaging # of avg

Real Time Automatic Automatic Enabled 32

Adjustment Procedure · Menu Selection Setting

Page 51 of the Agilent Technologies 1141A user manual — Attenuator Adapter Adjustment
Page 51 · click to enlarge
Page 51

Calibration Tests and Adjustment

Attenuator Adapter Adjustment

7 Connect the adapter/probe combination to the test board in the position shown in the figure below.

Signal to + input

8 Adjust the + HF RESP for best overall pulse response, the flattest pulse top. Use the figure below for adjust locations.

Agilent Technologies 1141A — Attenuator Adapter Adjustment diagram (page 51)
Fig. 25 · Attenuator Adapter Adjustment

9 Change the function generator to 10 Vp-p (10x adapter adjustment only). 10 On the oscilloscope, press CHAN and set the sensitivity to 1 mV/div.

Adjustment Locations

Page 52 of the Agilent Technologies 1141A user manual — Attenuator Adapter Adjustment
Page 52 · click to enlarge
Page 52

Calibration Tests and Adjustment Attenuator Adapter Adjustment

11 Connect the adapter/probe combination to the test board in the position shown in the figure below (signal to both inputs).

Signal to both input

12 Alternately adjust the LF CMRR and then the -HF RESP for a minimum signal on the oscilloscope. Repeat the adjustments until the signal is optimized to a minimum. Each adjustment should be set to minimize the component of the signal it affects most. Some high-frequency components of the signal are not affected by either adjustment.

Page 53 of the Agilent Technologies 1141A user manual — Service
Page 53 · click to enlarge
Page 53

Service

Page 54 of the Agilent Technologies 1141A user manual — Introduction
Page 54 · click to enlarge
Page 54

Introduction

This section provides troubleshooting, service, and repair information for the 1141A Differential Probe and 1142A Probe Control and Power Module. The troubleshooting information is provided to isolate a faulty assembly. When a faulty assembly has been located, the disassembly/assembly procedures help direct replacement of the assembly.

WAR N ING Maintenance should be performed by trained service personnel aware of the hazards involved (for example, fire and electric shock). When maintenance can be performed without power applied, the power cord must be removed from the instrument.

Service Introduction

Page 55 of the Agilent Technologies 1141A user manual — Performance Specifications and Characteristics
Page 55 · click to enlarge
Page 55

Service Performance Specifications and Characteristics

Performance Specifications and Characteristics

The following table gives performance specifications used to test the 1141A and 1142A. It also gives performance characteristics that are typical for the probe system.

Agilent Technologies 1141A — Performance Specifications and Characteristics diagram (page 55)
Fig. 26 · Performance Specifications and Characteristics

Parameter Probe alone With 10x attenuator With 100x attenuator

Bandwidth (-3 dB, dc coupled) dc to 200 MHz1

±200 V(dc + peak ac) ±500 V(dc + peak ac)

DC mode with no offset ± 300 mV peak ±3.0 V peak ±30 V peak

dc dc to 30 Hz 30 Hz to 200 MHz

dc Offset Range ±20 V ±200 V ±500 V

ac Low-freq. Response (-3dB) 15 Hz 1.5 Hz 1.5 Hz

dc Reject Response 5 Hz, 0.5 Hz, or 0.05 Hz (selectable irrespective of attenuator)

Displayed noises ≤ 50 µVrms Overload Recovery < 1 ms from overdrive that is less than the common mode range

Page 56 of the Agilent Technologies 1141A user manual — Performance Specifications and Characteristics
Page 56 · click to enlarge
Page 56

Service Performance Specifications and Characteristics

CMRR Specifications and Characteristics

A. CMRR specification for probe with no input adapters.

B. Low-frequency CMRR specification for probe with the ac adapter.

C. Typical CMRR characteristic for differential probe with no input adapters

D. Typical CMRR characteristic for differential probe with 100x attenuator adapter at input.

E. Typical CMRR characteristic for differential probe with 100x attenuator adapter at input.

Legend

Page 57 of the Agilent Technologies 1141A user manual — Performance Specifications and Characteristics
Page 57 · click to enlarge
Page 57
Agilent Technologies 1141A — Performance Specifications and Characteristics diagram (page 57)
Fig. 27 · Performance Specifications and Characteristics

Maximum Input Voltage vs. Frequency

A. Input voltage limits for probe alone.

B. Input voltage limits for 10x adapter.

C. Input voltage limits for 100x adapter.

Legend

Page 58 of the Agilent Technologies 1141A user manual — General Characteristics
Page 58 · click to enlarge
Page 58

General Characteristics

The following characteristics apply to the 1141A Differential Probe with the 1142A Probe Control and Power Module.

Temperature 0°C to +55 C° (32°F to +131°F) -40°C to +70°C (-40°F to +158°F)

Humidity up to 95% relative humidity (non- condensing) at +40°C (+104°F)

up to 90% relative humidity at +65°C (+149°F)

Altitude up to 4,600 meters (15,000 ft) up to 15,300 meters (50,000 ft)

Vibration Random vibration 5 to 500 Hz, 10 minutes per axis, 0.3grms.

Random vibration 5 to 500 Hz, 10 min. per axis, 2.41 grams. Resonant search 5 to 500 Hz swept sine, 1Octave/min. sweep rate, (0.75g), 5 min. resonant dwell at 4 resonances per axis.

Voltage90 to 132/198 to 264 Vac, 47 to 440 Hz Power: 25 VA maximum
Weight Netapproximately 1.8 kg (4.0 lb.) Shipping: approximately 2.7 kg (6.0 lb.)

Dimensions Refer to the outline drawings below.

Agilent Technologies 1141A — General Characteristics diagram (page 58)
Fig. 28 · General Characteristics
Show 5 more labels / short text on this page
Service General Characteristics · Environmental Conditions · Operating Non-operating · Power Requirements · Mechanical Dimensions
Page 59 of the Agilent Technologies 1141A user manual — Theory of Operation
Page 59 · click to enlarge
Page 59

Service Theory of Operation

Theory of Operation

The following discussion covers block-level theory for the 1141A/1142A differential probe system. Refer to the block diagram below. The differential probe system consists of two units, the 1141A Differential Probe with its accessories and the 1142A Probe Control and Power Module. For purposes of the following discussion, these will be called the probe and the control module respectively.

Agilent Technologies 1141A — Theory of Operation diagram (page 59)
Fig. 29 · Theory of Operation

Differential Probe System Block Diagram

Page 60 of the Agilent Technologies 1141A user manual — Control and Power Module
Page 60 · click to enlarge
Page 60

Service Theory of Operation

Differential Probe

The probe contains a two-path differential amplifier with unity gain. It is implemented on a double-sided surface-mount PC board with the high-frequency path on one side and the low-frequency path on the other. The two paths are split directly after the differential input connections.

High-Frequency Path The positive and negative inputs are ac-coupled at 33 Hz into identical impedance converters. The HF CMRR adjustment balances the gain at the outputs of the impedance converters. The impedance feed a semi-custom differential amplifier. An additional negative input to the differential amplifier brings in the sum of the low-frequency and feedback signals. The FREQ COMP adjustment provides variable high-frequency peaking of the differential amp. The output amp provides two signals. A feedback signal is summed with the low-frequency signals and the output signal is the final output of the probe. Overall probe gain is set by the HIGH FREQ GAIN adjustment. The entire signal portion of the high-frequency path is implemented on a hybrid IC. Support circuitry includes bias for the impedance converters and a bias supply for current sources on the hybrid.

Low-Frequency Path The bandwidth of the low-frequency path is approximately 75 KHz. The positive and negative inputs are dc coupled into identical inverting op-amps with gain of 0.5. They provide a precision 1 M Ω input impedance for the probe. The LOW FREQ BANDWIDTH adjustments match the gain and phase of the low-frequency path to that of the feedback from the probe output. One of the two adjustments is set to match the properties of the feedback and the other is to match the two low-frequency paths. These adjustments affect the CMRR quality of the probe. The inputs of the inverting op-amps include protection for ESD and over-voltage conditions. The inverting op-amps feed a precision differential amplifier with unity gain and a single- ended output. The output is fed to the summing amp and to the control module for use in the dc reject circuit.

Summing Amp The summing amp combines the feedback signal, the low-frequency signal, and the offset signal. The LOW FREQ GAIN adjustment matches the gain of the low-frequency path to the overall gain.

Control and Power Module

The control and power module provides offset functions, local and remote control, and power to the probe system.

Page 61 of the Agilent Technologies 1141A user manual — Attenuator Adapters
Page 61 · click to enlarge
Page 61

Service Theory of Operation

Offset Functions There are two offset functions developed in the control module: variable offset and dc reject. A variable offset voltage with coarse and fine adjustments can be selected by the front panel controls. The offset level is buffered by U8 and selected by multiplexer U3 as the input to offset amp U7. The output of the offset amp is summed with the low- frequency signal and feedback which gives dc coupling in the probe. Front panel screwdriver adjustment Offset Null zeros the dc output from the probe when the dc input and offset are zero. For dc Reject, an output from the low-frequency amplifier in the probe (LFSIG) is used to develop a voltage used to null the dc component of the input signal. LFSIG is an input to U6, an inverting amplifier and low-pass filter. Multiplexer U3 selects one of three capacitors to set a roll-off frequency of 0.05, 0.5, or 5 Hz. The output of U6 is selected, again by U3, as the input to the offset amp. When the output of the offset amp is summed into the low frequency path, the result is cancellation of the dc component of the input signal. Front panel screwdriver adjustment DC Reject Gain adjusts the gain of the dc reject circuit.

Local and Remote Control The front panel switch controls the dc reject and offset functions. It also selects remote operation, which allows control through the rear panel remote input connector.

Power Supply The supply provides ±6 V and ±15 V for the probe and analog control circuitry as well as +5 V for the digital control circuitry.

Attenuator Adapters

The 10x and 100x Attenuator Adapter are similar. A ceramic substrate carries two attenuators, one for each input polarity. A variable resistor adjusts the low-frequency balance (LF CMRR) between the two attenuators. The high-frequency adjustments are the same for each attenuator. Each attenuator is adjusted differently. One attenuator is adjusted for optimum pulse response and the other for best high-frequency CMRR.

Agilent Technologies 1141A — Attenuator Adapters diagram (page 61)
Fig. 30 · Attenuator Adapters

Attenuator Adapter, Simplified Schematic

Page 62 of the Agilent Technologies 1141A user manual — Test Board
Page 62 · click to enlarge
Page 62

Service Theory of Operation

Test Board

The test board is a device for conveniently connecting test signals to the differential probe. The probe can be connected to the board with the signal to the positive, negative, or both inputs.

Agilent Technologies 1141A — Test Board diagram (page 62)
Fig. 31 · Test Board

Test Board Schematic

Page 63 of the Agilent Technologies 1141A user manual — Service Policy
Page 63 · click to enlarge
Page 63

Service Policy

For parts of the 1141A/1142A probe system that are complex, the service policy is for assembly-level repair. For parts of the system with simple circuitry, the service policy is component-level repair. The service policy for the 1141A Differential Probe is assembly-level repair. Assemblies include the PC assembly and cable. The PC assembly is an “exchange assembly.” A repaired and tested assembly is shipped upon receipt of the defective assembly. The attenuator and ac coupling adapters are shipped as complete assemblies. The service policy for the 1142A Probe Control and Power Module is component-level repair.

Service Service Policy

Page 64 of the Agilent Technologies 1141A user manual — Probe Control and Power Module Troubleshooting
Page 64 · click to enlarge
Page 64

Troubleshooting

Use the following paragraphs to assist in troubleshooting problems with the 1141A/1142A Differential Probe.

Probe Troubleshooting

To troubleshoot the probe: 1 Apply a known signal to the input of the probe. 2 Check for an identical output at the output coax to the cable. This connection is the one soldered to the PC board. If the probe output cable is not terminated, or the coax is open, the output signal will be about twice the amplitude of the input signal. 3 If the signal is incorrect, check the power supply voltages from the 1142A Probe Control and Power Module. Use the cable diagram on the next page. 4 Troubleshoot the cable with an ohmmeter. Use the cable diagram on the next page.

Probe Control and Power Module Troubleshooting

The circuitry consists of simple power supplies, operational amplifiers, and TTL. Use conventional troubleshooting techniques. A complete parts list, component locator, and schematics are provided later in this chapter.

Service Troubleshooting

Page 65 of the Agilent Technologies 1141A user manual — Troubleshooting
Page 65 · click to enlarge
Page 65
Agilent Technologies 1141A — Troubleshooting diagram (page 65)
Fig. 32 · Troubleshooting

1142A Probe Control and Power Module

Service Troubleshooting

Page 66 of the Agilent Technologies 1141A user manual — Differential Probe
Page 66 · click to enlarge
Page 66

Service Removing and Replacing Assemblies

Removing and Replacing Assemblies

This section contains procedures for the removal and replacement of major assemblies.

Differential Probe

Use the following procedure to remove and replace the amplifier PC board in the differential probe.

Disassemble Probe 1 Remove the probe tip caps and probe tips. 2 Loosen the probe clamp ring at the cable end of the probe (1/4 turn counter- clockwise) and slide it down the cable. 3 Remove the bottom cover. a At the cable end of the probe, separate the covers about one centimeter (1/2 inch). b Slide the bottom cover toward the cable end of the probe until the locator pins at the probe input clear the holes. Then, remove the cover. Note the way the cable strain relief is keyed and held at the rear of the top cover. 4 Remove the probe top cover. The ground screw passes through the top cover and PC board and screws into the ground block.

a Remove the ground screw on the top of the probe. At the cable end, the PC board fits over the pins in the top cover.

b Lift the board off of the pins and slide it in the direction of the cable until the input connectors clear the front of the probe. 5 Un-solder the two connections where the coaxial output cable connects to the PC board. 6 Disconnect the cable connector from the probe PC board.

Page 67 of the Agilent Technologies 1141A user manual — Removing and Replacing Assemblies
Page 67 · click to enlarge
Page 67

Service Removing and Replacing Assemblies

Reassemble Probe The ground screw passes through the top cover and PC board and screws into the ground block. 1 If replacing the PC board, remove the input connectors from the old board and put them on the new one. 2 If replacing the cable, note the orientation of the probe clamp ring on the old cable, remove the ring and put it on the new cable. 3 Connect the cable connector to the PC board. 4 Solder the two connections of the coaxial cable to the PC board. On a new probe cable, the conductors of the coaxial cable are connected by a heavy single wire. Cut the heavy wire so it matches the wire on the cable that was removed. 5 Assemble the PC assembly into the top cover. The large hybrid is exposed when the assembly is in the top cover. The figure on the below shows the sequence of the ground screw, top cover, PC board, and ground block.

Agilent Technologies 1141A — Removing and Replacing Assemblies diagram (page 67)
Fig. 33 · Removing and Replacing Assemblies

a Insert the input connectors first, and seat the cable end of the PC assembly over the pins at the rear of the cover. b Position the ground block at the front of the PC assembly. c Insert the grounding screw through the top cover and screw it into the ground block as shown in the figure above. 6 Replace the bottom cover. a Position the cable strain relief and with one hand, hold the cable and top cover together. The flange on the strain relief has a notch that fits over a protrusion in the top cover.

Reassembling the Probe

Page 68 of the Agilent Technologies 1141A user manual — Probe Adapters
Page 68 · click to enlarge
Page 68
Agilent Technologies 1141A — Probe Adapters diagram (page 68)
Fig. 34 · Probe Adapters

b Insert the pins at the front of the bottom cover into the holes at the front of the top cover. c Close the covers together and fasten with the probe clamp ring. If the covers do not fit together tightly and easily, check for pinched wires (see caution above).

Probe Adapters

Use the following procedure to disassemble the probe adapters. The adapter housing consists of two plastic parts, one of which slides into the other. The parts are held together by the spring effect of two plastic tabs on the inner part. Mechanically, the ac adapter is about the same as the 10x and 100x attenuator adapters. The attenuator adapters have an extra ground connector which connects the substrate ground to the thumbwheel screw and plating inside the housing.

Disassemble Adapter 1 Remove the probe tip caps and probe tips from the adapter input. 2 Hold the adapter in one hand taking care not to block the output end of the adapter (the end which attaches to the probe). 3 Note the view of the input end of the adapter in the figure below. The arrows indicate the holding tabs.

4 With the thumb and forefinger, squeeze the tabs together, as indicated by the arrows. Simultaneously, push the tabs into the outer housing so the inner housing begins to slide out. 5 While holding the outer housing, push back against the thumbwheel until the inner housing can be grasped and removed.

Reassemble Adapter Reassembling the adapter is slightly harder because you have to align the connector pins and thumbwheel screw, while sliding the inner housing and outer housing together. 1 Be sure the input connectors and output pins are present and seated on the substrate or PC board.

Disassembling Adapters

Page 69 of the Agilent Technologies 1141A user manual — Probe Adapters
Page 69 · click to enlarge
Page 69

Service Removing and Replacing Assemblies

2 Combine the thumbwheel screw and thumbwheel and insert them into the hole in the outer housing. 3 Insert the substrate/board into the outer housing. Slip the attenuator ground (attenuator adapters only) over the thumbwheel screw and seat the input connectors in the proper holes in the housing.

Agilent Technologies 1141A — Probe Adapters diagram (page 69)
Fig. 35 · Probe Adapters

Do not force reassembly of the adapter. The housing halves will slide together with moderate friction. 4 Align the inner housing tabs with the grooves in the side of the outer housing and slide the two partly together. While seating the thumbwheel screw and pins, it will help to hold the assembly vertically, with the input end of the adapter down. This will allow the assemblies to sit vertically and more easily align with the holes in the inner housing. 5 Seat the thumbwheel screw first, then the output pins, into the appropriate holes in the inner housing. 6 Once the two housings are nearly together, press them together firmly until the tabs click into place.

Reassembling the Adapter

Page 70 of the Agilent Technologies 1141A user manual — Probe Control and Power Module
Page 70 · click to enlarge
Page 70

Service Removing and Replacing Assemblies

Probe Control and Power Module

Use the following procedure to disassemble the probe control and power module.

WAR N ING Hazardous voltages exist on the power supply. To avoid electrical shock, adhere closely to the following procedures.

1 Remove the power cord. 2 Remove four flathead screws and remove the top cover.

WAR N ING Be sure to reconnect the safety ground when reassembling the instrument.

  • 3 Unplug the safety ground from the tab on the rear panel of the instrument. 4 Note the orientation of the knobs. Remove the two knobs. 5 On the bottom of the instrument, remove the 5mm screw that fastens the transformer support. 6 Remove the following pan-head screws.
  • Three directly on the PC board.
  • Two on the ac input connector.
  • Two on the heatsink. 7 Remove the PC board. Slide it slightly forward so parts will clear the rear panel, then lift the rear of the board out while sliding it backwards. 8 Remove the two heatsink spacers from the standoffs that were directly under the heatsink. 9 Reverse the procedure to reassemble the control and power supply.
Page 71 of the Agilent Technologies 1141A user manual — Direct Mail Order System
Page 71 · click to enlarge
Page 71

Replaceable Parts

This section contains information for ordering parts. Service support for the 1141A Differential Probe is to the assembly level. Service support for the adapters is as complete assemblies, except for the probe tip caps and the probe tips. Service support for the 1142A Probe Control and Power Module is to the component level.

Parts List

  • The replaceable parts lists include all parts relevant to the applicable service levels. The information given for each part consists of the following:
  • Reference designator
  • Part number
  • Total quantity (Qty) in instrument or on assembly. The total quantity is given once and at the first appearance of the part number in the list.
  • Description of part
  • Typical manufacturer of part in a five-digit code.

Ordering Information

To order a part in the material part list, quote the part number, indicate the quantity desired, and address the order to the nearest Agilent Technologies Sales Office. To order a part not listed in the material list, include the instrument part number, instrument serial number, a description of the part (including its function), and the number of parts required. Address the order to the nearest Agilent Technologies Sales Office.

Direct Mail Order System

  • Within the USA, Agilent Technologies can supply parts through a direct mail order system. There are several advantages to this system:
  • Direct ordering and shipment from the Agilent Technologies parts center in California, USA.
  • No maximum or minimum on any mail order (there is a minimum amount for parts ordered through a local Agilent Technologies Sale Office when the orders require billing and invoicing).
  • Prepaid transportation (there is a small handling charge for each order).
  • No invoices. In order for Agilent Technologies to provide these advantages, a check or money order must accompany each order.

Service Replaceable Parts

Page 72 of the Agilent Technologies 1141A user manual — Manufacturers’ Code List
Page 72 · click to enlarge
Page 72

Mail order forms and specific ordering information are available through your local Agilent Technologies Sales Office. Addresses and telephone numbers are located in a separate document shipped with the manuals.

Manufacturers’ Codes

A list of manufacturers’ codes is given the table below. The codes are given for parts in the parts lists. The table gives the manufacturer and address for each code.

Manufacturers’ Code List

Mfr. No. Name Address

00000 Any satisfactory supplier

06665 Precision Monolithics Inc. Santa Clara, CA 95050

24546 Corning Glass Works (Bradford) Bradford, PA 16701

27014 National Semiconductor Corp Palo Alto, CA 94304

28480 Agilent Technologies Corporate Hq Palo Alto, CA 94304

32997 Bourns Inc. Riverside, CA 92507

Service Replaceable Parts

Page 73 of the Agilent Technologies 1141A user manual — Exploded View
Page 73 · click to enlarge
Page 73

Exploded View

Agilent Technologies 1141A — Exploded View diagram (page 73)
Fig. 36 · Exploded View

1141A Differential Probe Parts

Service Replaceable Parts

Page 74 of the Agilent Technologies 1141A user manual — Exploded View
Page 74 · click to enlarge
Page 74
Agilent Technologies 1141A — Exploded View diagram (page 74)
Fig. 37 · Exploded View

Power and Control Board Component Locator

Service Replaceable Parts

Page 75 of the Agilent Technologies 1141A user manual — Exploded View
Page 75 · click to enlarge
Page 75
Agilent Technologies 1141A — Exploded View diagram (page 75)
Fig. 38 · Exploded View

Service Replaceable Parts · Cabling Diagram

Page 76 of the Agilent Technologies 1141A user manual — Exploded View
Page 76 · click to enlarge
Page 76
Agilent Technologies 1141A — Exploded View diagram (page 76)
Fig. 39 · Exploded View

IC Connectors not Shown

Agilent Technologies 1141A — Exploded View diagram (page 76)
Fig. 40 · Exploded View

Supply Pin No. IC Group +15 14 U3 -15 3 GND 15 +15 7 U6-8, 11 -15 4 NC 1, 5, 8 +5 16 U9 GND 8

1142A Probe Control and Power Module

Page 77 of the Agilent Technologies 1141A user manual — Exploded View
Page 77 · click to enlarge
Page 77
Agilent Technologies 1141A — Exploded View diagram (page 77)
Fig. 41 · Exploded View

1142A Probe Control and Power Module

Page 78 of the Agilent Technologies 1141A user manual — MP5 01141-94307 1 DIFFERENTIAL PROBE LABEL 28480 01141-94301
Page 78 · click to enlarge
Page 78

1141A and 1142A Replaceable Parts

Ref. Des.

Qty Description Mfr. Code

Mfr. Part

A1 01141-66503 1 PC ASSEMBLY-DIFFERENTIAL PROBE 28480 01141-66501

E1 01141-26102 2 PROBING PIN 28480 01141-26102

E2 01141-22301 1 GROUND BLACK 28480 01141-22301

E3 01141-22401 1 GROUND SCREW 28480 01141-22401

MP1 01141-24102 2 CAP-PROBING PIN 28480 01141-24101

MP2 01141-44702 1 PROBE BOTTOM COVER 28480 01141-44701

MP3 01141-44102 1 PROBE TOP COVER 28480 01141-44101

MP4 01141-22502 1 PROBE POS CLAMP RING 28480 01141-22501

MP5 01141-94307 1 DIFFERENTIAL PROBE LABEL 28480 01141-94301

W1 01141-61603 1 PROBE CABLE ASSEMBLY 28480 01141-61601

01141-26102 2 PROBING PIN 28480 01141-26102

01141-24102 2 CAP-PROBING PIN 28480 01141-24101

01141-26102 2 PROBING PIN 28480 01141-26102

01141-24102 2 CAP-PROBING PIN 28480 01141-24101

01141-26102 2 PROBING PIN 28480 01141-26102

01141-24102 2 CAP-PROBING PIN 28480 01141-24101

1142A PROBE CONTROL AND POWER MODULE

Show 7 more labels / short text on this page
Service Replaceable Parts · Part Number · Number · 1141A Differential Probe · 10X Attenuator Adapter · 100X Attenuator Adapter · AC Adapter
Page 79 of the Agilent Technologies 1141A user manual — MP1 01142-47702 1 BOTTOM CHASSIS 28480 01142-47701
Page 79 · click to enlarge
Page 79

1141A and 1142A Replaceable Parts

Ref. Des.

Qty Description Mfr. Code

Mfr. Part

A1 01142-66501 1 PC ASSEMBLY-POWER AND CONTROL 28480 01142-66501

H1 0515-0374 7 SCREW-MACHINE M3 10mm-LG 00000 ORDER BY DESP.

H2 0515-1031 4 SCREW-MACHINE M3 6mm-LG 90-DEG-FLH-HD 00000 ORDER BY DESP.

H3 0515-1579 1 SCREW-MACHINE M5 18mm-LG 00000 ORDER BY DESP.

MP1 01142-47702 1 BOTTOM CHASSIS 28480 01142-47701

MP2 01142-44101 1 TOP CHASSIS 28480 01142-44101

MP3 01142-24701 2 SPACER-HEAT SINK 28480 01142-24701

MP4 0370-1097 2 KNOB-POINTER 28480 0340-1097

MP5 0403-1012 4 FOOT 28480 0403-0727

MP6 5041-0234 5 KEYCAP 28480 5041-0234

W1 8120-1521 1 POWER CORD 18-AWG 3-COND 90-IN-LG (US/ Canada only)

Show 4 more labels / short text on this page
Service Replaceable Parts · Part Number · Number · 28480 8120-1521
Page 80 of the Agilent Technologies 1141A user manual — C21-22 NOT ASSIGNED
Page 80 · click to enlarge
Page 80

Power Supply Replaceable Parts

Ref. Des. Part Number

Qty Description Mfr. Code

Mfr. Part

Prefix the reference designators with A1

C1 0160-6190 2 CAPACITOR-FXD 0.33UF ±10% 50VDC 28480 0160-6190

C2 0160-5474 1 CAPACITOR-FXD 0.1UF ±5% 100VDC MET-POLY 28480 0160-5474

C3 0180-3298 2 CAPACITOR-FXD 2200UF+30-10% 50VDC AL 28480 0180-3298

C4 0180-3298 CAPACITOR-FXD 2200UF+30-10% 50VDC AL 28480 0180-3298

C5 0160-6500 4 CAPACITOR-FXD 0.01UF ±10% 100VDC CER 28480 0160-6500

C6 0160-6500 CAPACITOR-FXD 0.01UF ±10% 100VDC CER 28480 0160-6500

C7 0160-6500 CAPACITOR-FXD 0.01UF ±10% 100VDC CER 28480 0160-6500

C8 0160-6500 CAPACITOR-FXD 0.01UF ±10% 100VDC CER 28480 0160-6500

C9 0180-3845 4 CAPACITOR-FXD 4.7UF ±10% 35VDC TA 28480 0180-3845

C10 0180-3845 CAPACITOR-FXD 4.7UF ±10% 35VDC TA 28480 0180-3845

C11 NOT ASSIGNED

C12 0160-5581 1 CAPACITOR-FXD 0.033UF ±10% 63VDC 28480 0160-5581

C13 0160-6190 CAPACITOR-FXD 0.33UF ±10% 50VDC 28480 0160-6190

C14 0160-7060 1 CAPACITOR-FXD 3.3UF ±10% 63VDC 28480 0160-7060

C15 0160-4801 1 CAPACITOR-FXD 100PF ±5% 100VDC CER 28480 0160-4801

C16 0180-3845 CAPACITOR-FXD 4.7UF ±10% 35VDC TA 28480 0180-3845

C17 0180-3845 CAPACITOR-FXD 3.3UF ±10% 63VDC 28480 0180-3845

C18 NOT ASSIGNED

C19 0160-5471 1 CAPACITOR-FXD 0.1UF ±5% 50VDC MET-POLY 28480 0160-5471

C20 0160-5469 1 CAPACITOR-FXD 1UF ±10% 50VDC MET-POLY 28480 0160-5469

C21-22 NOT ASSIGNED

C23 0180-3784 4 CAPACITOR-FXD 22UF ±20% 25VDC TA 28480 0180-3784

C24 0180-3784 CAPACITOR-FXD 22UF ±20% 25VDC TA 28480 0180-3784

C25 0180-3784 CAPACITOR-FXD 22UF ±20% 25VDC TA 28480 0180-3784

C26 0180-3784 CAPACITOR-FXD 22UF ±20% 25VDC TA 28480 0180-3784

Service Replaceable Parts · Number

Page 81 of the Agilent Technologies 1141A user manual — HS1 01142-21101 1 HEAT SINK 28480 01142-21101
Page 81 · click to enlarge
Page 81

Power Supply Replaceable Parts

Ref. Des. Part Number

Qty Description Mfr. Code

Mfr. Part

CR1 NOT ASSIGNED

CR2 1901-1087 2 DIODE-PWR RECT 100V 3A 04713 MR501

CR3 1901-1087 DIODE-PWR RECT 100V 3A 04713 MR501

CR4 NOT ASSIGNED

CR5 1901-0731 8 DIODE-PWR RECT 400V 1A 28480 1901-0731

CR6 1901-0731 8 DIODE-PWR RECT 400V 1A 28480 1901-0731

CR7-8 NOT ASSIGNED

CR9 1901-0731 8 DIODE-PWR RECT 400V 1A 28480 1901-0731

CR10 1901-0731 8 DIODE-PWR RECT 400V 1A 28480 1901-0731

CR11 1901-0731 8 DIODE-PWR RECT 400V 1A 28480 1901-0731

CR12 1901-0731 8 DIODE-PWR RECT 400V 1A 28480 1901-0731

CR13 1901-0734 2 DIODE-PWR RECT 1N5818 30V 1A 04713 IN5818

CR14 1901-0734 2 DIODE-PWR RECT 1N5818 30V 1A 04713 IN5818

CR15-16 NOT ASSIGNED

CR17 1901-0731 DIODE-PWR RECT 400V 1A 28480 1901-0731

CR18 1901-0731 DIODE-PWR RECT 400V 1A 28480 1901-0731

DS1 1990-0521 1 LED-LAMP LUM-INT=2.2MCD IF=50MA-MAX 24840 5082-4955

E1 2110-0642 1 FUSEHOLDER 6.3A 250A 28480 2110-0642

E2 2110-0565 1 FUSEHOLDER CAP 28480 2110-0565

F1 2110-0201 1 FUSE 0.25A 250V TD 28480 2110-0201

H1 0515-1579 1 SCREW-MACHINE M5 18mm-LG 28480 0515-1579

HS1 01142-21101 1 HEAT SINK 28480 01142-21101

Service Replaceable Parts · Number

Page 82 of the Agilent Technologies 1141A user manual — MP8 01142-24702 1 TRANSFORMER SUPPORT 28480 01142-24702
Page 82 · click to enlarge
Page 82

Power Supply Replaceable Parts

Ref. Des. Part Number

Qty Description Mfr. Code

Mfr. Part

J1 1252-4731 1 CONNECTOR-AC PWR 28480 1251-4743

J2 1252-1487 1 CONN-RECT D-SUBMIN 9-CKT (remote) 28480 1252-1487

J3 1252-3935 1 CONNECTOR-ROUND 8-CKT (probe) 28480 1252-3134

MP1 1400-1604 1 LED MOUNT 28480 1400-1604

MP2 1205-0732 4 SPRING CLIP 28480 1205-0732

MP3 0361-0685 3 RIVET-BLIND DR-PIN RNDH 0.125DIA 28480 0361-0685

MP4 0340-1211 3 INSULATOR-THERMAL 28480 0340-1211

MP5-6 NOT ASSIGNED

MP7 1400-0249 1 CABLE TIE 0.062-0.625-DIA 0.091-WD NYL 16956 08-465/GRY

MP8 01142-24702 1 TRANSFORMER SUPPORT 28480 01142-24702

MP9 01142-28801 1 WASHER-TRANSFORMER SUPPORT 28480 01142-28801

R1-2 NOT ASSIGNED

R3 0757-0442 4 RESISTOR 10K 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-1002-F

R4 0757-0465 1 RESISTOR 100K 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-1003-F

R5 0757-0199 5 RESISTOR 21.5K 1% 0.125W TC=0±100 24546 CT4-1/8-TO-2152-F

R6 0757-0199 RESISTOR 21.5K 1% 0.125W TC=0±100 24546 CT4-1/8-TO-2152-F

R7 0698-4431 2 RESISTOR 2.05K 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-2052-F

R8 0699-1203 4 RESISTOR 120.0 1% 0.125W TF TC=0±25 28480 0699-1203

R9 0757-0434 2 RESISTOR 3.65K 1% 0.125W TF TC=±100 24546 CT4-1/8-TO-3651-F

R10 0698-4431 RESISTOR 2.05K 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-2052-F

R11 0699-1203 RESISTOR 120.0 1% 0.125W TF TC=0±25 28480 0699-1203

R12 0757-0434 RESISTOR 3.65K 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-3651-F

R13 0757-0420 1 RESISTOR 750 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-751-F

R14 0698-4002 2 RESISTOR 5K 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-5001-F

R15 0699-1203 RESISTOR 120.0 1% 0.125W TF TC=0±25 28480 0699-1203

Service Replaceable Parts · Number

Page 83 of the Agilent Technologies 1141A user manual — R41 2100-4250 2 RESISTOR-VAR 10K 20% 28480 2100-4250
Page 83 · click to enlarge
Page 83

Power Supply Replaceable Parts

Ref. Des. Part Number

Qty Description Mfr. Code

Mfr. Part

R16 0698-6317 2 RESISTOR 500 0.1% 0.125W TF TC=0±25 28480 0698-6317

R17 0698-4002 RESISTOR 5K 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-5001-F

R18 0699-1203 RESISTOR 120.0 1% 0.125W TF TC=0±25 28480 0699-1203

R19 0698-6317 RESISTOR 500 0.1% 0.125W TF TC=0±25 28480 0698-6317

R20 NOT ASSIGNED

R21 0698-8827 1 RESISTOR 1M 1% 0.125W TF TC=0±100 28480 0698-8827

R22 0757-0442 RESISTOR 10K 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-1002-F

R23 0757-0427 1 RESISTOR 1.5K 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-1501-F

R24 0757-0401 1 RESISTOR 100 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-101-F

R25 2100-3161 1 RESISTOR-TRMR 20K 10% TKF SIDE-ADJ 32997 3006P-1-203

R26 2100-3056 1 RESISTOR-TRMR 5K 10% TKF SIDE-ADJ 17-TRN 32997 3006P-1-502

R27 0757-0199 RESISTOR 21.5K 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-2152-F

R28 0757-0199 RESISTOR 21.5K 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-2152-F

R29 0757-0442 RESISTOR 10K 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-1002-F

R30 0757-0442 RESISTOR 10K 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-1002-F

R30 0757-0280 1 RESISTOR 1K 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-1001-F

R32 0757-0460 1 RESISTOR 61.9K 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-6192-F

R33-39 NOT ASSIGNED

R40 0757-0123 1 RESISTOR 34.8K 1% 0.125W TF TC=0±100 28480 0757-0123

R41 2100-4250 2 RESISTOR-VAR 10K 20% 28480 2100-4250

R42 2100-4250 RESISTOR-VAR 10K 20% 28480 2100-4250

R43 0757-0458 1 RESISTOR 51.1K 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-5112-F

R44 0683-2755 1 RESISTOR 2.7M 1% 0.25W TF TC=0±100 28480 0683-2755

R45 0698-4517 1 RESISTOR 127K 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-1273-F

R46 0698-3271 1 RESISTOR 115K 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-1153-F

R47 0698-8961 1 RESISTOR 909K 1% 0.125W TF TC=0±100 28480 0698-8961

R48 0757-0199 RESISTOR 21.5K 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-2152-F

Service Replaceable Parts · Number

Page 84 of the Agilent Technologies 1141A user manual — S2 3101-3007 1 SWITCH-6 STATION ASSEMBLY 28480 3101-3007
Page 84 · click to enlarge
Page 84

Power Supply Replaceable Parts

Ref. Des. Part Number

Qty Description Mfr. Code

Mfr. Part

RP1 1810-1242 1 RESISTOR NETWORK 28480 1810-1242

S1 3101-2609 1 SWITCH-SL DPST STD 5A 250VAC PC 28480 3101-2609

S2 3101-3007 1 SWITCH-6 STATION ASSEMBLY 28480 3101-3007

T1 9100-4750 1 TRANSFORMER-POWER (with mtg. hardware) 28480 9100-4750

U1 1826-1403 2 IC V RGLTER-ADJ-POS 3/40V 00000 LT317AT

U2 1826-1670 2 IC V RGLTR-ADJ-NEG -37/1.2V TO-220 PKG 00000 LT337AT

U3 5081-9233 1 ANALOG MULTIPLEXER 4 CHNL 16 -DIP-P 28480 1820-2182

U4 1826-1403 IC V RGLTR ADJ-POS 3/40V 00000 LT317AT

U5 1826-1670 IC V RGLTR-ADJ-NEG -37/1.2V TO-220 PKG 00000 LT337AT

U6 1826-1381 3 IC OP AMP LOW-BIAS-H-IMPD 8-DIP-P PKG 00000 LT1012CN8

U7 1826-1381 IC OP AMP LOW-BIAS-H-IMPD 8-DIP-P PKG 00000 LT1012CN8

U8 1826-1381 IC OP AMP LOW-BIAS-H-IMPD 8-DIP-P PKG 00000 LT1012CN8

U9 1820-3177 1 IC MUXR/DATA-SEL CMOS/74HC 2-TO-1-LINE 04713 MC74HC157N

U10 1826-0774 1 IC V RGLTR-V-REF-FXD 1.22/1.24V TO-92 27014 LM385BZ-1.2

U1 1826-0635 1 IC OP AMP LOW-OFS 8-DIP-P PKG 06665 OP-07CP

VR1 1902-0951 1 DIODE-ZNR 5.1V 5% D0-35 PD=.4W TC±.035% 28480 1902-0951

W1 01141-61602 1 CABLE ASSEMBLY, SAFETY GROUND 28480 01141-61602

Service Replaceable Parts · Number

Page 85 of the Agilent Technologies 1141A user manual — Locator Table for Control and Power Supply
Page 85 · click to enlarge
Page 85

Locator Table for Control and Power Supply

Ref. Des.

C1 C-2 C17 C-6 CR10 C-5 J1 G-7 R5 B-6 R22 B-3 R45 C-3 U5 D-6

C2 F-3 C18 B-5 CR11 C-3 J2 B-7 R6 B-6 R23 B-5 R46 C-3 U6 B-3

C3 H-2 C19 B-6 CR12 C-6 J3 C-7 R7 C-4 R24 B-5 R47 C-2 U7 C-3

C4 J-3 C20 B-5 CR13 C-3 R8 C-5 R25 B-2 R48 D-3 U8 B-3

C5 C-3 C21 B-5 CR14 C-6 MP1 H-1 R9 C-4 R26 B-2 U9 A-5

C6 C-5 C22 B-6 CR17 B-5 MP2 E-3 R10 C-5 R27 A-5 RP1 B-3 U10 C-2

C7 B-3 C23 B-4 CR18 B-6 MP3 D-3 R11 C-5 R28 A-5 U11 C-3

C8 C-5 C24 B-6 MP4 E-5 R12 C-5 R29 B-6 S1 H-7

C9 C-5 C25 B-4 DS1 G-2 MP7 E-4 R13 E-3 R30 B-6 S2 E-2 VR1 E-3

C10 C-5 C26 G-6 MP8 G-4 R14 C-4 R31 B-3

C11 E-3 E1 F-6 MP9 G-4 R15 C-4 R32 B-2 T1 H-5

C12 A-3 CR2 F-3 E2 E-6 R16 C-4 R40 C-2

C13 A-3 CR3 F-3 P1 E-7 R17 A-2 R41 A-2 U1 D-4

C14 B-4 CR5 C-3 F1 E-6 R18 C-6 R42 C-2 U2 D-5

C15 A-3 CR6 C-5 R3 A-6 R19 C-6 R43 C-2 U3 A-2

C16 C-4 CR9 C-3 HS1 D-5 R4 A-6 R21 B-3 R44 C-3 U4 D-3

Show 24 more labels / short text on this page
Service Replaceable Parts · Grid · Ref Des · Grid · Ref Des · Grid · Ref Des · Grid · Ref Des · Grid · Ref Des · Grid · Ref Des · Grid · Ref Des · Grid · Loc · Loc · Loc · Loc · Loc · Loc · Loc · Loc
Page 89 of the Agilent Technologies 1141A user manual — Agilent Technologies P.O. Box 2197 1900 Garden of the Gods Road
Page 89 · click to enlarge
Page 89

Notices

Warranty The material contained in this document is provided “as is,” and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control.

commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR 52.227- 14 (June 1987) or DFAR 252.227- 7015 (b)(2) (November 1995), as applicable in any technical data.

© Agilent Technologies, Inc. 2000-2004

No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws.

Manual Safety Notices CAUTION:

Manual Part Number 01141-97002, July 2004

A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met. WARNING:

01141-97000, June 2000 01141-97001, September 2002 01141-97002, July 2004

Agilent Technologies, Inc. 1900 Garden of the Gods Road Colorado Springs, CO 80907 USA

A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.

Technology Licenses The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.

Restricted Rights Legend If software is for use in the performance of a U.S. Government prime contract or subcontract, Software is delivered and licensed as “Commercial computer software” as defined in DFAR 252.227-7014 (June 1995), or as a “commercial item” as defined in FAR 2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard

Agilent Technologies P.O. Box 2197 1900 Garden of the Gods Road

Print History

Page 90 of the Agilent Technologies 1141A user manual — !
Page 90 · click to enlarge
Page 90

Product Safety Notices

any unintended operation.

Service instructions are for trained service personnel. To avoid dangerous electric shock, do not perform any service unless qualified to do so. Do not attempt internal service or adjustment unless another person, capable of rendering first aid and resuscitation, is present.

!

This apparatus has been designed and tested in accordance with IEC Publication 1010, Safety Requirements for Measuring Apparatus, and has been supplied in a safe condition. This is a Safety Class I instrument (provided with terminal for protective earthing). Before applying power, verify that the correct safety precautions are taken (see the following warnings). In addition, note the external markings on the instrument that are described under "Safety Symbols."

Instruction manual symbol: the product is marked with this symbol when it is necessary for you to refer to the instruction manual in order to protect against damage to the product.

Do not install substitute parts or perform any unauthorized modification to the instrument.

Capacitors inside the instrument may retain a charge even if the instrument is disconnected from its source of supply.

Hazardous voltage symbol.

Earth terminal symbol: Used to indicate a circuit common connected to grounded chassis.

Do not operate the instrument in the presence of flammable gasses or fumes. Operation of any electrical instrument in such an environment constitutes a definite safety hazard.

Warnings Before turning on the instrument, you must connect the protective earth terminal of the instrument to the protective conductor of the (mains) power cord. The mains plug shall only be inserted in a socket outlet provided with a protective earth contact. You must not negate the protective action by using an extension cord (power cable) without a protective conductor (grounding). Grounding one conductor of a two-conductor outlet is not sufficient protection.

Do not use the instrument in a manner not specified by the manufacturer.

To clean the instrument If the instrument requires cleaning: (1) Remove power from the instrument. (2) Clean the external surfaces of the instrument with a soft cloth dampened with a mixture of mild detergent and water. (3) Make sure that the instrument is completely dry before reconnecting it to a power source.

Only fuses with the required rated current, voltage, and specified type (normal blow, time delay, etc.) should be used. Do not use repaired fuses or short-circuited fuseholders. To do so could cause a shock or fire hazard.

If you energize this instrument by an auto transformer (for voltage reduction or mains isolation), the common terminal must be connected to the earth terminal of the power source.

Whenever it is likely that the ground protection is impaired, you must make the instrument inoperative and secure it against

Safety Symbols

Ask about the Agilent Technologies 1141A

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the Agilent Technologies 1141A

Be the first to ask — other owners or our team will answer.

Original document

Agilent Technologies 1141A manual PDF

90 pages·10.1 MB·Thermometer·PDF
Click any page above to open it here
Page 1 / …
100%
Loading pages…