Publication number 01141-97002 July 2004
For Safety and Regulatory information, see the pages at the end of this book.
© Copyright Agilent Technologies 2000-2004
All Rights Reserved

This manual contains information for use and service of the differential probe system, the 1141A Differential Probe and 1142A Probe Control and Power Module. In this document, the two models will be treated as a system.
Each of the two instrument models that make up the differential probe system has a serial number sticker. The sticker for the 1141A Differential Probe is inside the probe, in the bottom cover. (See chapter 3 for disassembly procedure.)
The 1141A/1142A probe system allows measurement of small differential signals in the presence of much larger common-mode signals. It has the following major features:
The variable offset capability can be used to measure small ac signals in the presence of much larger dc levels. Remote operation of key features allows the use of the probe system in automatic test situations.
The probe can be used with an oscilloscope, spectrum analyzer, or any instrument where differential probing is required and a compatible 50 Ω input is available.



Operating the Probe Accessories Supplied
This chapter shows you how to connect and operate the 1141A Differential Probe and 1142A Probe Control and Power Module as a differential probe system.
The following items are supplied as part of the 1141A/1142A probe system. Item numbers refer to the numbers in Figure 1-1 on page -7 and Figure 1-2 on page -8. Those without item numbers are supplied but not shown in figures. See the Replaceable Parts List for parts not listed below.

Operating the Probe Accessories Supplied

1141A Differential Probe and Accessories

Operating the Probe Accessories Available


1142A Probe Control and Power Module
The following accessories can be ordered.
1141A Miscellaneous Accessories

Operating the Probe To inspect the probe
Inspect the shipping container for damage. If the shipping container or cushioning material is damaged, it should be kept until the contents of the shipment have been checked for completeness and the instrument had been checked mechanically and electrically. Accessories supplied with the instrument are listed in see “Accessories Supplied” on page 6 of this manual. If the contents are incomplete, if there is mechanical damage or defect, or if the instrument does not pass calibration tests, notify the nearest Agilent Technologies office. If the shipping container is damaged, or the cushioning materials show sign of stress, notify the carrier as well as the nearest Agilent Technologies office. Keep the shipping materials for the carrier’s inspection. The office will arrange for repair or replacement at Agilent Technologies’ option without waiting for a claim settlement.
The 1141A/1142A probe system can be used with other instruments as well as oscilloscopes. You can use it with a spectrum analyzer or frequency counter, or any instrument with an input that can be terminated with 50 Ω. If you are going to use the probe system with an instrument other than an oscilloscope, you may need to set up the probe with an oscilloscope first. This will allow you to select coupling and reject modes, and set offset, so the output of the probe is compatible with signal requirements of the other instrument.
The following table is a list of the test equipment required to test calibration, make adjustments, and troubleshoot this instrument. The table indicates the critical specifications of the test equipment and for which procedure the equipment is necessary. Equipment other than the recommended model may be used if it satisfies the critical specifications listed in the table.
Recommended Agilent Model/Part Use
Equipment Required Critical Specifications
Signal Generator 1 to 200 MHz, 400 m Vrms 8648A P
Function Generator 2.5 to 500 kHz, 600 m Vp-p to 16 Vp-p 33120A A
Oscilloscope >400 MHz bandwidth, 1 mV/div sensitivity 54830B P,A
dc source ± 100 mV to 7 Vdc E3632A P
Recommended Test Equipment

Operating the Probe Recommended Test Equipment
Recommended Agilent Model/Part Use
Equipment Required Critical Specifications
Test Board No substitute (supplied accessory) 01141-66504 A
Load BNC Feedthrough, 50 Ω Pasternack Enterprises PE6008- 50 or Huber+Suhner 22543742
Cables (2) BNC, 50 Ω 36-inch 10503A P,A
Cable BNC, 50 Ω 9-inch 10502A A
Cable Type N (m) 24-inch 11500B P
Adapter Type N (m) to BNC (f) 1250-0780 A
Adapter Type N (f) to BNC (m) 1250-0077 P
Adapters (2) BNC (f) to dual banana (m) 1251-2277 P
Alignment tool Small flat blade (supplied accessory) 8710-1961/ Sprague-Goodman part number GTT-5G
Setting up the probe
The following paragraphs cover system preparation and initial adjustments.
Power Requirements The 1141A/1142A probe system (specifically the 1142A) requires a power source of either 90 to 132/198 to 264 Vac, 47 to 440 Hz, 25 VA maximum.
Line Voltage Selection Before applying power, verify the setting of the LINE SELECT switch on the rear panel of the 1142A. The slide switch can be set to either 115 or 230 V.
P=Calibration Tests, A=Adjustments


WAR N ING Before connecting this instrument, the protective earth terminal of the instrument must be connected to the protective conductor of the (Mains) power cord. The Mains plug must be inserted in a socket outlet provided with a protective each contact. The protective action must be negated by the use on an extension cord (power cable) without a protective conductor (grounding). Grounding one conductor of a two- conductor outlet does not provide an instrument ground.
This instrument is provided with a three-wire power cable. When connected to an appropriate ac power outlet, this cable grounds the instrument cabinet. The type of power cable plug shipped with the instrument depends on the country of destination. The 1142A Power Control and Power Module does not have a power switch. A power switch is not required because of the low mains power requirement.

Procedure 1 Use the power cord to connect the 1142A to the ac mains. 2 Connect the 1141A probe cable power connector to the PROBE connector on the rear panel of the 1142A power module. 3 Connect the output of the probe to the input of the oscilloscope. 4 Set the input impedance of the oscilloscope to 50 Ω. If the oscilloscope does not have a selectable 50 Ω input impedance, connect a 50 Ω BNC feedthrough termination between the probe output and the input of the oscilloscope. 5 If making an initial equipment setup, continue with the initial adjustment in the following section.
For a given combination of 1141A Differential Probe and 1142A Probe Control and Power Module, you may want to adjust the Offset Null and DC Reject Gain. Typically, you need to make these adjustments only once, before the probe is first used. You can make them any time to optimize the system. These adjustments do not affect the specifications of the probe system.
1142A Rear Panel

Operating the Probe Recommended Test Equipment

Equipment Setup Use the following procedure to setup the differential probe system for initial adjustment.
1 Use the probe setup procedure to set up the probe system. 2 Connect a 50 Ω BNC feedthrough terminator to the output of the probe. 3 Disconnect all accessories from the input of the probe. 4 Connect the DVM to measure the dc output of the probe at the 50 Ω load.
1142A Front Panel

Operating the Probe Recommended Test Equipment
5 Set up the 1142A: a Set the Local/Remote push button to Local. b Under DC Couple, press Zero offset. 6 Set the power supply output to 5 V. 7 Arrange a connection between the power supply and the test board. The negative terminal of the supply should connect to the shield of the test board BNC. If your power supply has standard binding posts, you can connect a banana-to-BNC adapter to the supply and connect a BNC cable between the supply and the test board.
Adjustment Warm up the 1141A for 30 minutes before making adjustments. 1 With the 1141A probe inputs unconnected, adjust Offset Null on the 1142A for a minimum reading on the DVM. The voltage swing of the adjustment is approximately ±4 mV. 2 On the 1142A, under DC Reject, press 5.0 Hz. 3 Read and record the reading on the DVM, mV. 4 Connect the probe to the test board in the position shown below (signal to + input).
Signal to + input
5 After the DVM reading stabilizes, adjust DC Reject Gain to the reading recorded in step 3. With a 5 V supply, the voltage swing is approximately ±12.5 mV. With a lower supply, the voltage swings proportionally less.
The 1141A Differential Probe and accessories are designed to provide a variety of ways to connect to circuitry and make measurements. In the descriptions, any method used to connect to the probe signal inputs also applied to the adapters. The figure below shows, in a general way, the use of accessories.

Operating the Probe Recommended Test Equipment


Probe Tips Probe tips fit into the receptacles in the probe and are held in place with probe tip caps. If necessary, you can solder the probe tips into a circuit or wires can be soldered to the tips. If you solder to the probe tips, be careful not to melt the plastic probe tip caps.
No te Because of the close tolerances between the probe tip caps and probe tips, it will be difficult to separate the probe tips and caps once the probe tips have been soldered.
Ground Leads The circular end of the ground lead fits over the screw on the top side of the probe.
Basic Accessory Connections

Operating the Probe Recommended Test Equipment
No te Use extension leads and similar connection accessories carefully. Extension leads compromise the high-frequency specifications of the probe. CMRR is particularly sensitive to unbalanced input parameters.
Mini-Grabbers Mini-grabbers can be attached to the probe or adapter through the extension leads. 1 Remove the probe tip caps and tips. 2 Attach the extension leads to the probe or adapter. 3 Attach the mini-grabbers to the extension leads.
Circuit Connector Posts These 0.025-inch square posts can be used to connect either directly to the probe or to the extension leads. Solder the posts directly into your circuitry or use them to make extension leads that plug into the inputs of the probe or adapters.
Shielded Signal Leads The shielded signal leads allow connection to points in a circuit that are up to 10 inches apart. The leads are shielded to within approximately 1/2 inch of the end of the lead so they minimize pick-up due to stray fields from adjacent circuitry. 1 Connect the end with the ground connector to the probe pins and ground of the differential probe or adapter. 2 Connect the free ends of the leads to 0.025-inch square or 0.030-inch round pins in your circuitry or to the mini-grabbers.
No te Each lead has an input capacitance of approximately 15 pF. This capacitance may limit the bandwidth of your measurement (depending on the impedance of the circuit). Also, CMRR may be affected because of slight differences between the input capacitance of the two leads. CMRR is also affected by differences in impedance between the two measurements points.

Operating the Probe Recommended Test Equipment
Adapters There are three adapters for use with the differential probe. Two adapters are attenuators, a 10x and a100x. The other is an ac adapter for blocking dc from the probe input. The adapters are installed on the probe after the probe tip caps and probe tips have been removed. The adapter fastens to the probe using a thumb wheel located on the underside of the adapter. the figure below shows a good way to hold the probe while attaching the adapter. 1 Remove the probe tip caps and probe tips, from the probe. 2 Fit the adapter over the end of the probe and rotate the thumb wheel with your finger until the adapter fits snugly. A snug fit is important because the ground is maintained through the thumb wheel screw. A loosely attached adapter compromises the mechanical and electrical integrity of the combination.


Operating the Probe Recommended Test Equipment


Allowed Adapter Connections

Operating the Probe Recommended Test Equipment
Test Board The primary use of the test board is to apply test and calibration signals to the input of the probe or adapters. Specific use of the test board is covered wherever it applies.
Grounding is very important when probing circuitry. Improper grounding can increase the common mode signal level. This reduces the effectiveness of the differential probe. The mechanical connections at the input of the probe are ground for probe signals. The screw where the ground lead attaches (see figure 1-7) fastens to this ground. Also, the attenuator and ac adapter fasten to this ground through the screw connection and the ground is carried through each adapter to its front.

There are three methods for blocking or compensating for the dc component of a signal. Each has specific advantages.
Probe Grounds

Operating the Probe Recommended Test Equipment
Probe System Coupling Functions
dc offset dc Reject ac Coupling Adapter
Probe alone Probe with 10x Probe with 100x
± 20 V ± 200 V ± 500 V
± 20 V ± 200 V ± 500 V
± 20 V ± 500 V ± 500 V
Set-up needed Adjust offset to put signal on screen
Select DC Reject low-frequency corner
Attach ac Coupling adapter to differential probe
Remote Control? Yes1 Yes1 No
1 Isolated external dc reference and control signals are needed

Operating the Probe Recommended Test Equipment
To use dc reject: 1 Remove the ac adapter if it is installed. 2 On the front panel of the 1142A, press Local. 3 Under DC Reject on the front panel, press 5.0 Hz or 0.5 Hz individually, or 5.0 Hz and 0.5 Hz simultaneously to get 0.05 Hz. Within the frequency and voltage characteristics noted elsewhere in this manual, low frequencies are nulled from the input signal.

Operating the Probe Recommended Test Equipment
To use ac coupling: 1 Attach the ac coupling adapter to the input of the probe or the input of the attenuator adapter. 2 On the 1142A, press Local and Zero offset.
For automatic test applications, the coupling and offset functions provided by the 1142A Probe Control and Power Module can be remotely controlled through a connector on the rear panel of the module. The connection is through a standard 9-pin female D-subminiature connector. This style is the same as that used on some personal computer monitor cables, which provides an economical way to connect the 1142A to the controller interface on an automatic test system. The following table gives the connections.

1 Function Select 1 (A1R) 6 Function Select 0 (A0R)
NOT E To minimize dc offset errors and potential noise coupling, electrically isolate all connections between the Remote Input connector and the controlling system.

Operating the Probe Recommended Test Equipment
Function Select The easiest way to control the function select lines is contact closures between the lines and Digital common (pin 2) of the remote input connector. (TTL compatible control signals can be used; but to avoid problems with ground loops, they must be electrically isolated.) The following truth table shows the functions provided by the function select lines. For the Remote Inputs, “0” represents a closure and “1” represents an open circuit.
Remote Functions Select Truth Table
0 0 0.05 Hz Reject
0 1 0.5 Hz Reject
1 0 5 Hz Reject
1 1 DC Couple
Variable Offset The remote variable offset can be used when the dc couple function is remotely selected. The offset voltage must be referenced to the External offset common (pin 4) of the remote input connector. It must be electrically isolated from the controlling system. The following table shows the offset range and remote offset requirements for probe and adapter combinations.
Remote Offset Input Requirements
Offset Range Remote Requirements
Probe alone ± 20 V ± 10 V
Probe with 10x adapter ± 200 V ± 10 V
Probe with 100x adapter ± 500 V ± 2.5 V
If the instrument requires cleaning: (1) Remove power from the instrument. (2) Clean the external surfaces of the instrument with a soft cloth dampened with a mixture of mild detergent and water. (3) Make sure that the instrument is completely dry before reconnecting it to a power source.
Remote Input Function · A1R (Pin 1) · A0R (Pin 6)

Operating the Probe Recommended Test Equipment

Operating the Probe Recommended Test Equipment


Calibration Tests and Adjustment Equipment Required
This chapter is divided into two sections. The first section gives calibration tests and the second adjustment procedures for the 1141A Differential Probe and 1142A Probe Control and Power Module.
A complete list of equipment required for the calibration tests and adjustments is listed in “Recommended Test Equipment” on page 9. Equipment required for individual procedures is listed at the procedure. Any equipment satisfying the critical specifications listed may be substituted for the recommended model.
The test board is a supplied accessory for use during calibration tests and adjustments to connect signals to the differential probe (with or without adapters). A BNC connector connects the test board to a cable from the signal generator. The board includes a 50 Ω termination (two 100 Ω resistors).
Once the probe tip caps and probe tips have been removed, the probe can be connected to the test board in one of three ways, as shown in the figure below. At each position of the probe, probe inputs are connected to a different combination of signal and ground. A separate terminal on the test board connects the ground of the probe to the signal ground.
Test Board Showing Probe Positions

Calibration Tests and Adjustment
Calibration Tests
These procedures test the probe’s electrical performance using applicable specifications given in “Performance Specifications and Characteristics” on page 55 as performance standards. Specifications applicable to individual tests are noted at the test for reference.
Testing Interval The calibration testing procedures may be performed for incoming inspection of the instrument and should be performed periodically thereafter to ensure and maintain peak performance. The recommended test interval is yearly or every 2,000 hours of operation. Amount of use, environmental conditions, and the user’s experience concerning need for testing will contribute to verification requirements.
Calibration Test Record The results of the calibration tests may be tabulated in the Calibration Test Record provided at the end of the calibration tests. The Calibration Test Record listed the calibration tests and provides an area to mark test results. The results recorded in the Calibration Test Record during initial inspection may be used for later comparisons of the tests during periodic maintenance, troubleshooting, and after repairs or adjustments.
Calibration Test Procedures Procedures may be done individually or in any order.
NOT E Allow the instrument to warm up for at least 30 minutes prior to beginning calibration tests.

Calibration Tests and Adjustment dc Gain Accuracy
This test checks the dc gain accuracy of the differential probe and the dc accuracy of the differential probe with attenuator and adapters.
Equipment Required Critical Specifications
ac/dc Calibrator or dc Power Supply
100 mV to 7 V
100 mV to 7 V E3632A
DVM 0.5% accuracy, 10 µV resolution 3458A or E34401A
Load BNC Feedthrough, 50 Ω Pasternack Enterprises PE6008-50 or Huber+Suhner 22543742
Cables (2) BNC 50 Ω 10503A
Adapters (2) BNC (f) to dual banana (m) 1251-2277
Test Board No substitute 01141-66514
Probe Gain Test Procedure 1 Connect the probe and test equipment as shown in figure 2-2.
Equipment Required · Recommended Agilent Model/Part

Calibration Tests and Adjustment

2 Set up the 1142A probe control and power module as follows: a Set the Local/Remote push button to Local. b Under DC Couple, press the Zero offset button. 3 With the 1141A Probe Amp disconnected from the test PCA, adjust the Offset Null control on the 1142A until the DVM reads 0Vdc. If the probe output voltage cannot be set to 0V, subtract this voltage from the subsequent measurements in this test. 4 Connect the input of the Probe Amp to the test board in the position shown in Figure 2-4 on page -31. 5 Adjust the DC Source to output 100 mV (nom.) 6 Record the Vin1 measurement from the top DVM in figure 2-2. 7 Record the Vout1 measurement from the bottom DVM in figure 2-2. 8 Connect the Probe Amp to measure a negative voltage as shown in figure 2-3. 9 Record the Vout2 measurement from the bottom DVM in figure 2-2


Signal to - input
10 Calculate probe gain as
∆Vin Vout1 Vout2 – ( )
Record the result of this calculation in the “Calibration Test Record” on page 39. To pass this test, the probe gain = 0.98 to 1.02
NOT E Failure of the gain accuracy test can be caused by mis-adjustment of the probe. Perform the Probe Adjustment procedure in the Adjustments section later in this chapter and retest.
10x Attenuator Accuracy Test
NOT E If the gain test for the probe fails, the 10x Attenuator Accuracy Test will fail or the results will be poor. Do not continue until the probe passes the dc gain test.
1 Disconnect the probe from the test board and connect the 10x attenuator adapter to the probe. 2 Carefully connect the input of the probe/attenuator to the test board in the position shown in the figure below (signal to + input).


Signal to + input
3 Set the dc calibrator output 3 V dc. 4 Record the Vin1 measurement from the top DVM in figure 2-2. 5 Record the Vout1 measurement from the bottom DVM in figure 2-2 6 Carefully connect the input of the probe/attenuator to the test board in the position shown in the figure below (signal to - input).
Signal to - input
7 Record the Vout2 measurement from the bottom DVM in figure 2-2. 8 Calculate the 10x attenuator gain as
∆Vin Vout1 Vout2 – ( )
Record the result of this calculation in the “Calibration Test Record” on page 39.
NOT E Failure of the accuracy test for the 10x attenuator can be caused by mis-adjustment of the low-frequency CMRR (LF CMRR) adjustment. Perform the Attenuator Adapter Adjustment procedure in the Adjustments section later in this chapter then retest the attenuator adapter. If if continues to fail, repair is necessary.

Calibration Tests and Adjustment dc Gain Accuracy
100x Attenuator Accuracy Test
NOT E If the gain test for the probe fails, it will be reflected in the test for the 100x attenuator adapter. Do not continue until the probe passes the gain test.
1 Disconnect the probe/attenuator from the test board. Remove the 10x attenuator adapter from the probe and connect the 100x attenuator adapter. 2 Carefully connect the input of the probe/attenuator to the test board in the position shown in the figure below (signal to + input).
Signal to + input
3 Set the dc source output to 7 Vdc. 4 Record the Vin1 measurement from the top DVM in figure 2-2. 5 Record the Vout1 measurement from the bottom DVM in figure 2-2

Calibration Tests and Adjustment
6 Carefully connect the input of the probe/attenuator to the test board in the position shown in the figure below (signal to - input).
Signal to - input
7 Record the Vout2 measurement from the bottom DVM in figure 2-2. 8 Calculate the 100x attenuator gain as
∆Vin Vout1 Vout2 – ( )
2 V in 1 × --------------------------------- =
Record the result of this calculation in the “Calibration Test Record” on page 39
NOT E Failure of the accuracy test for the 100x attenuator can be caused by mis-adjustment of the low-frequency CMRR (LF CMRR) adjustment. Perform the Attenuator Adapter Adjustment procedure in the Adjustments section later in this chapter then retest the attenuator adapter. If if continues to fail, repair is necessary.

Calibration Tests and Adjustment Bandwidth
This test checks the high-frequency response of the 1141A Differential Probe. The bandwidth of the oscilloscope is characterized first so it is not a factor in the measurement.
Critical Specifications Recommended Model/Part
Oscilloscope 400 MHz bandwidth 54830A
Signal Generator 200 MHz at ≈ 230 mVrms 8648A
Test Board No substitute 01141-66504
Cable Type N (m) 24-inch 11500B
Adapter Type N (f) to BNC (m) 1250-0077
Procedure This test depends on the accuracy of the termination on the test board and the termination in the oscilloscope. Both should be with 1%. 1 With the N cable and N-to-BNC adapter, connect the signal generator to the oscilloscope channel 1 input. 2 Set the signal generator for 200 MHz at 0 dBm (about 224 mVrms). 3 Set the 1142A front panel switches to Local and Zero offset. 4 On the oscilloscope press AUTOSCALE, then set the following parameters.
TIMEBASE (time/div) 2 ns/div
CHAN 1 (sensitivity) (input R)
100 mV/div 50 Ω DC
ACQUISITION Sampling Mode Memory Depth Sample Rate Averaging # of avg
Real Time Automatic Automatic Enabled 32
5 The signal on screen should be about six divisions amplitude. Measure the peak-to-peak voltage of the channel 1 signal and record the reading. Vp-p (1) = mV
Equipment Required · Equipment Required · Menu Selection Setting

Calibration Tests and Adjustment
6 Reconfigure the equipment. a Disconnect the signal generator cable from the oscilloscope input and connect it to the test board. b Connect the output of the differential probe to the channel 1 input of the oscilloscope. c Carefully connect the input of the probe to the test board in the position shown in the figure below (signal to + input).
Signal to + input
7 Record the Vp-p reading on the oscilloscope. Vp-p (1) = mV 8 Divide the reading from step 6 by the reading from step 4.
Answer from step 6 Answer from step 4 =
Record the result in the Calibration Test Record. The result should be 0.707 or greater, indicating a probe bandwidth of 200 MHz or more. The bandwidth can be checked at other signal levels. Change the signal generator output level and oscilloscope V/div range proportionally.
NOT E Failure of the bandwidth test can be caused by mis-adjustment of the probe. Perform the Probe Adjustment procedure in the Adjustments section later in this chapter.

Calibration Tests and Adjustment CMRR Test
This test checks the CMRR at 1 MHz and 100 MHz.
Specification 3000:1 at 1 MHz, 10:1 at 100 MHz

Signal Generator 1-100 MHz at = 400 mVrms 8648A
Procedure 1 Connect the probe power connector to the PROBE connection on the rear of the 1142A Probe Control and Power Module. 2 Set the 1142A front panel switches to Local and Zero offset. 3 Connect the probe output to the oscilloscope channel 1 input. 4 Connect the input of the probe to the test board in the position shown in the figure below (signal to + input).
Signal to + input
5 Connect the signal generator to the test board. 6 Set the signal generator for 1 MHz at 385 mVrms (1 Vp-p).

Calibration Tests and Adjustment
7 On the oscilloscope, press AUTOSCALE and set the following parameters.

ACQUISITION Sampling Mode Memory Depth Sample Rate Averaging # of avg
8 On the oscilloscope, measure the peak-to-peak voltage of the channel 1 signal, then V P-P, then press 1) and record the reading. Vp-p (1) = mV 9 Connect the input of the probe to the test board in the position shown in the figure below (signal to both inputs).
Signal to both input
10 Set the sensitivity to 1 mV/div. 11 After the measurement settles (averaging is complete), record the V P-P reading. Vp-p (2) = mV 12 Disconnect the probe amp from the test board and measure Vnoise pp on channel 1. 13 Calculate the CMRR result as follows
Vpp1 Vpp2 Vnoisepp – =
14 The result in step 12 should be ≥ 3000, representing a CMRR of 3000:1 or more. Record the CMRR in the Calibration Test Record.

Calibration Tests and Adjustment CMRR Test
15 Connect the input of the probe to the test board in the position shown in the figure below (signal to + input).
Signal to + input
16 Set the signal generator for 100 MHz at 0.0 dBm (about 224 mVrms, 632 mVp-p). 17 Set the oscilloscope to channel 1 and change the horizontal scale to 5 ns/div. 18 After the measurement settles (averaging is complete), note the V P-P reading. Vp-p (1) = mV 19 Connect the input of the probe to the test board in the position shown in the figure below (signal to both inputs).
Signal to both input
20 Set the channel 1 sensitivity to 10 mV/div. 21 After the measurement settles (averaging is complete), not the V P-P reading. Vp-p (2) = mV 22 Disconnect the probe amp from the test board and measure Vnoise pp on channel 1. 23 Calculate the CMRR result as follows
Vpp1 Vpp2 Vnoisepp – =
24 The result in step 21 should be ≥ 10, representing a CMRR of 10:1 or more. Record the CMRR in the Calibration Test Record.

Calibration Tests and Adjustment
Calibration Test Record
1141A/1142A Differential Probe Tested by
Serial No Work Order No
Recommended Test Interval - 1 Year/2000 hours Date
Recommended next testing Temperature
dc Gain Accuracy Probe Only
+0.98 mV to +1.02 mV
10x +0.096 mV to +0.104 mV
100x +0.0096 mV to 0.0104 mV
Bandwidth >0.707 at 200 MHz
CMRR 1 MHz ≥3000:1
100 MHz ≥10:1

Calibration Tests and Adjustment Probe Adjustment
This section provides adjustment procedures for the 1141A Differential Probe and attenuator adapters. There are no service adjustments for the 1142A Probe Control Module.
Adjustment Interval None of the adjustment procedures that follow should be considered for a routine maintenance plan. The differential probe and attenuator adapters should be adjusted under conditions specified at the beginning of the respective procedures.
NOT E Warm up the instrument for 30 minutes before starting adjustment procedures.
This procedure adjust the high-frequency and low-frequency paths on the 1141A Differential Probe.
NOT E Do not perform this procedure as a part of routine maintenance. Perform the procedure only if the probe does not meet specifications or has been repaired.
CAUTION You are going to remove the covers of the probe, so the assembly inside will be exposed while under power. The PC assembly will be electrically and mechanically vulnerable. Do these adjustment procedures in an ESD-safe area. Avoid inadvertent contact between the powered assembly and nearby tools and equipment. Avoid mechanical damage by carefully handling the exposed assembly and cables.

Calibration Tests and Adjustment
The following equipment is required for this procedure. Procedures are based on the model or part number recommended.
Equipment Required Critical Specifications Recommended Agilent Model/Part
Function Generator 2.5 kHz, 1 Vp-p 33120A
Oscilloscope 300 MHz at 1 mV/div 54830B
Signal Generator 200 MHz at ≈ 300 mVrms 8648A
Test Board No substitute 01141-66504
Cables (2) BNC, 50 Ω 10503A
Adapter Type N (m) to BNC (f) 1250-0780
Probe Preparation The probe cover must be removed before adjustment. Drift due to temperature differences with and without covers is negligible. 1 Remove the probe tip caps and probe tips. 2 Loosen the probe clamp ring at the cable end of the probe (1/4 turn counter- clockwise) and slide it down the cable. 3 Remove the bottom cover. a At the cable end of the probe, separate the covers about centimeter (1/2 inch). b Slide the bottom cover toward the cable end of the probe until the locator pins at the probe input clear the holes. Then, remove the cover.
NOT E Note the position of the ground block at the input end of the probe. The ground block is held, through the PC assembly, by the grounding screw on the top of the probe. The ground block must be reinstalled on the PC assembly after the top cover is removed. Handle the PC assembly by the edges of the PC board.
4 Remove the ground connection screw on the top of the probe. The ground block will become free. At the cable end, the PC board fits over pins inside the top cover. 5 Lift the board off of the pins in the cover and slide it in the direction of the cable until the input connectors clear the front of the probe.
Equipment Required

Calibration Tests and Adjustment Adjustment Procedure
6 As shown in the figure below, use the grounding screw to reinstall the ground block on the PC assembly.

Attaching Ground Block to Probe PC Assembly
The ground block provides a mechanical and electrical connection when the probe PC assembly is connected to the test board. 7 Connect the probe power connector to the PROBE connection on the rear of the 1142A Probe Control and Power Module. 8 Connect the mains power to the 1142A. 9 Set the 1142A front panel switches to Local and Zero offset.
Unless specified elsewhere, the procedures must be followed in the order given. The only adjustment which may be done separately is HF COMP, the high-frequency compensation.

Calibration Tests and Adjustment
3 On the oscilloscope, then press CLEAR DISPLAY. Press AUTOSCALE, then set up the following parameters.

ACQUISITION Sampling Mode Memory Depth Sample Rate Averaging # of avg
4 On the oscilloscope, measure the peak-to-peak voltage of the channel 1 signal and record the reading. Vp-p (1) = mV 5 Disconnect the BNC cable from the channel 1 input and connect it to the BNC connector on the test board. 6 Connect the output of the probe to the channel 1 input. 7 Carefully connect the input of the probe to the test board in the position shown in the figure below (signal to + input).
Signal to + input

Calibration Tests and Adjustment Adjustment Procedure
8 Center adjustment R11, HF CMRR (see following figure).

R11, HF CMRR Adjustment
9 Adjust R9, HF GAIN so the Vp-p (1) measurement is the same as in step 4, within ±0.5%. Make the adjustment slowly so the oscilloscope display has time to react to signal averaging. Press CLEAR DISPLAY occasionally to restart averaging, which gives a quicker indication of changes. 10 Carefully connect the probe to the test board in the position shown in the figure below (signal to both inputs).
Signal to both inputs
11 Set the function generator output to 1.0 Vp-p. 12 On the oscilloscope, set the channel 1 sensitivity to 1.00 mV/div. 13 Adjust R11 for minimum signal amplitude as shown in Vp-p reading for channel 1. Adjust R11 slowly and use CLEAR DISPLAY frequently to restart averaging.

Calibration Tests and Adjustment
2 Change the oscilloscope settings to:
TIMEBASE (time/div) 50 µs/div
CHAN 1 (sensitivity) 100 mV/div
3 Carefully connect the input of the probe to the test board in the position shown in the figure below (signal to + input).
Signal to + input
4 Adjust R14 (LF Gain) and C4 (+ LF BANDWIDTH) for the flattest pulse top (see figure below). Again, adjust slowly and press CLEAR DISPLAY frequently to restart averaging.

R14 and C4 Adjustment
Menu Selection Setting

Calibration Tests and Adjustment Adjustment Procedure
5 Carefully connect the probe to the test board in position shown in the figure below (signal to both inputs).
Signal to both inputs
TIMEBASE (time/div) 50 µs/div
CHAN 1 (sensitivity) 2 mV/div
8 Adjust C6 (-LF BANDWIDTH) for minimum signal amplitude on the oscilloscope. Again, adjust C6 slowly and press CLEAR DISPLAY frequently to restart averaging.
Menu Selection Setting

Calibration Tests and Adjustment
High Frequency Compensation This adjustment sequence continues from the Low Frequency Response and CMRR adjustment. However, it can be done separately if the probe meets all specifications except bandwidth. Adjust R13 for unity gain at 200 MHz. 1 Connect the signal generator to the test board and set it for 200 MHz and 300mVp-p (107 mVrms). 2 Carefully connect the input of the probe to the test board in the position shown in the figure below (signal to + input).
Signal to + input
3 Press AUTOSCALE, then measure the peak-to-peak voltage on channel 1 (Press SHIFT (blue), press V P-P, then press 1). 4 Adjust R13 (HF COMP) to make the signal amplitude measurement on the oscilloscope 300 mVp-p, or as close to that as possible. Adjust slowly and press CLEAR DISPLAY frequently to restart averaging.

The minimum allowable amplitude is 212 mVp-p. Typical values will be between 275 and 325 mVp-p. the probe needs repair if the minimum cannot be reached.
Probe Reassembly 1 Disconnect the probe power cable at the rear panel of the 1142A 2 Remove the probe PC assembly from the test board. Be sure the probe input connectors remain attached to the probe. 3 Remove the grounding screw and ground block from the PC assembly.
R13 Adjustment

Calibration Tests and Adjustment Adjustment Procedure
4 Assemble the PC assembly in the top cover. The side of the PC assembly with the large hybrid is exposed when the assembly is in the top cover. The figure below shows how the top cover, PC board, and ground block fit together.

a Insert the input connectors first, and seat the cable end of the PC assembly over the pins at the rear of the cover. b Position the ground block at the center-front of the PC assembly. c Insert the grounding screw through the top cover and PC assembly and screw it into the ground block as shown. 5 Replace the bottom cover. a Position the cable strain relief and with one hand, hold the cable and top together. The flange on the strain relief has a notch that fits around a protrusion in the top cover.
b Insert the pins at the front of the bottom cover into the holes at the front of the top cover. c Close the two covers together and fasten with the probe clamp ring.
Reassembling the Probe

Calibration Tests and Adjustment
Attenuator Adapter Adjustment
The following procedure should be used if it is necessary to adjust an attenuator adapter. Attenuator adapters have only characteristics; they do not have any specifications. An adapter will need adjustment only if one or more of the following occurs.
NOT E Attenuator adapters should not be adjusted as part of routine maintenance. The design necessary to give the high CMRR and high impedance of the attenuators makes adjustment delicate. Additionally, because of broadband noise, a spectrum analyzer should be used to verify adjustment quality.
Each attenuator contains two identical voltage dividers on a ceramic substrate. There are three adjustments. The low-frequency CMRR adjust the low-frequency balance between the two voltage dividers. The two high-frequency adjustments are electrically identical. Each adjusts the high-frequency compensation of one of the voltage dividers. To meet both pulse response and CMRR characteristics they are adjusted differently. Briefly, the adjustment procedure is:
a Adjust the positive high-frequency response (+ HF RESP) for the best pulse response using a 3.5 kHz square wave. b Adjust the negative high-frequency response (-HF RESP) and Low-frequency CMRR (LF CMRR) for best CMRR using a 3.5 kHz square wave. The following equipment is required for this procedure. Procedures are based on the model or part number recommended.
Equipment Required Critical Specifications Recommended Model/Part
Function Generator 3.5 kHz, 3 V to 16 Vp-p 33120A
Oscilloscope 100 MHz at 1 mV/div 54830B
Probe/Power Module No substitute 1141A/1142A
Test Board No substitute 01141-66504
Cable BNC, 50 Ω 9-inch 10502A
Cable BNC, 50 Ω 36-inch 10503A
Alignment tool Flat blade (supplied accessory) 8710-1961/ Sprague-Goodman part, GTT-5G
Equipment Required

Calibration Tests and Adjustment Attenuator Adapter Adjustment
NOT E The attenuator must be adjusted when installed on the 1141A probe with which it will be used. The specifications and characteristics will not be met if the attenuator adapter is adjusted with one differential probe then used with another.
TIMEBASE (time/div) 50 µs/div
CHAN 1 (sensitivity) (input R)
50 mV/div 50 Ω DC
TRIG (mode) source level
trg’d EXT 1.00000 V
ACQUISITION Sampling Mode Memory Depth Sample Rate Averaging # of avg
Real Time Automatic Automatic Enabled 32
Adjustment Procedure · Menu Selection Setting

Calibration Tests and Adjustment
7 Connect the adapter/probe combination to the test board in the position shown in the figure below.
Signal to + input
8 Adjust the + HF RESP for best overall pulse response, the flattest pulse top. Use the figure below for adjust locations.

9 Change the function generator to 10 Vp-p (10x adapter adjustment only). 10 On the oscilloscope, press CHAN and set the sensitivity to 1 mV/div.
Adjustment Locations

Calibration Tests and Adjustment Attenuator Adapter Adjustment
11 Connect the adapter/probe combination to the test board in the position shown in the figure below (signal to both inputs).
Signal to both input
12 Alternately adjust the LF CMRR and then the -HF RESP for a minimum signal on the oscilloscope. Repeat the adjustments until the signal is optimized to a minimum. Each adjustment should be set to minimize the component of the signal it affects most. Some high-frequency components of the signal are not affected by either adjustment.


This section provides troubleshooting, service, and repair information for the 1141A Differential Probe and 1142A Probe Control and Power Module. The troubleshooting information is provided to isolate a faulty assembly. When a faulty assembly has been located, the disassembly/assembly procedures help direct replacement of the assembly.
WAR N ING Maintenance should be performed by trained service personnel aware of the hazards involved (for example, fire and electric shock). When maintenance can be performed without power applied, the power cord must be removed from the instrument.
Service Introduction

Service Performance Specifications and Characteristics
The following table gives performance specifications used to test the 1141A and 1142A. It also gives performance characteristics that are typical for the probe system.

Parameter Probe alone With 10x attenuator With 100x attenuator
Bandwidth (-3 dB, dc coupled) dc to 200 MHz1
±200 V(dc + peak ac) ±500 V(dc + peak ac)
DC mode with no offset ± 300 mV peak ±3.0 V peak ±30 V peak
dc dc to 30 Hz 30 Hz to 200 MHz
dc Offset Range ±20 V ±200 V ±500 V
ac Low-freq. Response (-3dB) 15 Hz 1.5 Hz 1.5 Hz
dc Reject Response 5 Hz, 0.5 Hz, or 0.05 Hz (selectable irrespective of attenuator)
Displayed noises ≤ 50 µVrms Overload Recovery < 1 ms from overdrive that is less than the common mode range

Service Performance Specifications and Characteristics
CMRR Specifications and Characteristics
A. CMRR specification for probe with no input adapters.
B. Low-frequency CMRR specification for probe with the ac adapter.
C. Typical CMRR characteristic for differential probe with no input adapters
D. Typical CMRR characteristic for differential probe with 100x attenuator adapter at input.
E. Typical CMRR characteristic for differential probe with 100x attenuator adapter at input.
Legend


Maximum Input Voltage vs. Frequency
A. Input voltage limits for probe alone.
B. Input voltage limits for 10x adapter.
C. Input voltage limits for 100x adapter.
Legend

The following characteristics apply to the 1141A Differential Probe with the 1142A Probe Control and Power Module.
Temperature 0°C to +55 C° (32°F to +131°F) -40°C to +70°C (-40°F to +158°F)
Humidity up to 95% relative humidity (non- condensing) at +40°C (+104°F)
up to 90% relative humidity at +65°C (+149°F)
Altitude up to 4,600 meters (15,000 ft) up to 15,300 meters (50,000 ft)
Vibration Random vibration 5 to 500 Hz, 10 minutes per axis, 0.3grms.
Random vibration 5 to 500 Hz, 10 min. per axis, 2.41 grams. Resonant search 5 to 500 Hz swept sine, 1Octave/min. sweep rate, (0.75g), 5 min. resonant dwell at 4 resonances per axis.
Dimensions Refer to the outline drawings below.


Service Theory of Operation
The following discussion covers block-level theory for the 1141A/1142A differential probe system. Refer to the block diagram below. The differential probe system consists of two units, the 1141A Differential Probe with its accessories and the 1142A Probe Control and Power Module. For purposes of the following discussion, these will be called the probe and the control module respectively.

Differential Probe System Block Diagram

Service Theory of Operation
The probe contains a two-path differential amplifier with unity gain. It is implemented on a double-sided surface-mount PC board with the high-frequency path on one side and the low-frequency path on the other. The two paths are split directly after the differential input connections.
High-Frequency Path The positive and negative inputs are ac-coupled at 33 Hz into identical impedance converters. The HF CMRR adjustment balances the gain at the outputs of the impedance converters. The impedance feed a semi-custom differential amplifier. An additional negative input to the differential amplifier brings in the sum of the low-frequency and feedback signals. The FREQ COMP adjustment provides variable high-frequency peaking of the differential amp. The output amp provides two signals. A feedback signal is summed with the low-frequency signals and the output signal is the final output of the probe. Overall probe gain is set by the HIGH FREQ GAIN adjustment. The entire signal portion of the high-frequency path is implemented on a hybrid IC. Support circuitry includes bias for the impedance converters and a bias supply for current sources on the hybrid.
Low-Frequency Path The bandwidth of the low-frequency path is approximately 75 KHz. The positive and negative inputs are dc coupled into identical inverting op-amps with gain of 0.5. They provide a precision 1 M Ω input impedance for the probe. The LOW FREQ BANDWIDTH adjustments match the gain and phase of the low-frequency path to that of the feedback from the probe output. One of the two adjustments is set to match the properties of the feedback and the other is to match the two low-frequency paths. These adjustments affect the CMRR quality of the probe. The inputs of the inverting op-amps include protection for ESD and over-voltage conditions. The inverting op-amps feed a precision differential amplifier with unity gain and a single- ended output. The output is fed to the summing amp and to the control module for use in the dc reject circuit.
Summing Amp The summing amp combines the feedback signal, the low-frequency signal, and the offset signal. The LOW FREQ GAIN adjustment matches the gain of the low-frequency path to the overall gain.
The control and power module provides offset functions, local and remote control, and power to the probe system.

Service Theory of Operation
Offset Functions There are two offset functions developed in the control module: variable offset and dc reject. A variable offset voltage with coarse and fine adjustments can be selected by the front panel controls. The offset level is buffered by U8 and selected by multiplexer U3 as the input to offset amp U7. The output of the offset amp is summed with the low- frequency signal and feedback which gives dc coupling in the probe. Front panel screwdriver adjustment Offset Null zeros the dc output from the probe when the dc input and offset are zero. For dc Reject, an output from the low-frequency amplifier in the probe (LFSIG) is used to develop a voltage used to null the dc component of the input signal. LFSIG is an input to U6, an inverting amplifier and low-pass filter. Multiplexer U3 selects one of three capacitors to set a roll-off frequency of 0.05, 0.5, or 5 Hz. The output of U6 is selected, again by U3, as the input to the offset amp. When the output of the offset amp is summed into the low frequency path, the result is cancellation of the dc component of the input signal. Front panel screwdriver adjustment DC Reject Gain adjusts the gain of the dc reject circuit.
Local and Remote Control The front panel switch controls the dc reject and offset functions. It also selects remote operation, which allows control through the rear panel remote input connector.
Power Supply The supply provides ±6 V and ±15 V for the probe and analog control circuitry as well as +5 V for the digital control circuitry.
The 10x and 100x Attenuator Adapter are similar. A ceramic substrate carries two attenuators, one for each input polarity. A variable resistor adjusts the low-frequency balance (LF CMRR) between the two attenuators. The high-frequency adjustments are the same for each attenuator. Each attenuator is adjusted differently. One attenuator is adjusted for optimum pulse response and the other for best high-frequency CMRR.

Attenuator Adapter, Simplified Schematic

Service Theory of Operation
The test board is a device for conveniently connecting test signals to the differential probe. The probe can be connected to the board with the signal to the positive, negative, or both inputs.

Test Board Schematic

For parts of the 1141A/1142A probe system that are complex, the service policy is for assembly-level repair. For parts of the system with simple circuitry, the service policy is component-level repair. The service policy for the 1141A Differential Probe is assembly-level repair. Assemblies include the PC assembly and cable. The PC assembly is an “exchange assembly.” A repaired and tested assembly is shipped upon receipt of the defective assembly. The attenuator and ac coupling adapters are shipped as complete assemblies. The service policy for the 1142A Probe Control and Power Module is component-level repair.
Service Service Policy

Use the following paragraphs to assist in troubleshooting problems with the 1141A/1142A Differential Probe.
To troubleshoot the probe: 1 Apply a known signal to the input of the probe. 2 Check for an identical output at the output coax to the cable. This connection is the one soldered to the PC board. If the probe output cable is not terminated, or the coax is open, the output signal will be about twice the amplitude of the input signal. 3 If the signal is incorrect, check the power supply voltages from the 1142A Probe Control and Power Module. Use the cable diagram on the next page. 4 Troubleshoot the cable with an ohmmeter. Use the cable diagram on the next page.
The circuitry consists of simple power supplies, operational amplifiers, and TTL. Use conventional troubleshooting techniques. A complete parts list, component locator, and schematics are provided later in this chapter.
Service Troubleshooting


1142A Probe Control and Power Module
Service Troubleshooting

Service Removing and Replacing Assemblies
This section contains procedures for the removal and replacement of major assemblies.
Use the following procedure to remove and replace the amplifier PC board in the differential probe.
Disassemble Probe 1 Remove the probe tip caps and probe tips. 2 Loosen the probe clamp ring at the cable end of the probe (1/4 turn counter- clockwise) and slide it down the cable. 3 Remove the bottom cover. a At the cable end of the probe, separate the covers about one centimeter (1/2 inch). b Slide the bottom cover toward the cable end of the probe until the locator pins at the probe input clear the holes. Then, remove the cover. Note the way the cable strain relief is keyed and held at the rear of the top cover. 4 Remove the probe top cover. The ground screw passes through the top cover and PC board and screws into the ground block.
a Remove the ground screw on the top of the probe. At the cable end, the PC board fits over the pins in the top cover.
b Lift the board off of the pins and slide it in the direction of the cable until the input connectors clear the front of the probe. 5 Un-solder the two connections where the coaxial output cable connects to the PC board. 6 Disconnect the cable connector from the probe PC board.

Service Removing and Replacing Assemblies
Reassemble Probe The ground screw passes through the top cover and PC board and screws into the ground block. 1 If replacing the PC board, remove the input connectors from the old board and put them on the new one. 2 If replacing the cable, note the orientation of the probe clamp ring on the old cable, remove the ring and put it on the new cable. 3 Connect the cable connector to the PC board. 4 Solder the two connections of the coaxial cable to the PC board. On a new probe cable, the conductors of the coaxial cable are connected by a heavy single wire. Cut the heavy wire so it matches the wire on the cable that was removed. 5 Assemble the PC assembly into the top cover. The large hybrid is exposed when the assembly is in the top cover. The figure on the below shows the sequence of the ground screw, top cover, PC board, and ground block.

a Insert the input connectors first, and seat the cable end of the PC assembly over the pins at the rear of the cover. b Position the ground block at the front of the PC assembly. c Insert the grounding screw through the top cover and screw it into the ground block as shown in the figure above. 6 Replace the bottom cover. a Position the cable strain relief and with one hand, hold the cable and top cover together. The flange on the strain relief has a notch that fits over a protrusion in the top cover.
Reassembling the Probe


b Insert the pins at the front of the bottom cover into the holes at the front of the top cover. c Close the covers together and fasten with the probe clamp ring. If the covers do not fit together tightly and easily, check for pinched wires (see caution above).
Use the following procedure to disassemble the probe adapters. The adapter housing consists of two plastic parts, one of which slides into the other. The parts are held together by the spring effect of two plastic tabs on the inner part. Mechanically, the ac adapter is about the same as the 10x and 100x attenuator adapters. The attenuator adapters have an extra ground connector which connects the substrate ground to the thumbwheel screw and plating inside the housing.
Disassemble Adapter 1 Remove the probe tip caps and probe tips from the adapter input. 2 Hold the adapter in one hand taking care not to block the output end of the adapter (the end which attaches to the probe). 3 Note the view of the input end of the adapter in the figure below. The arrows indicate the holding tabs.
4 With the thumb and forefinger, squeeze the tabs together, as indicated by the arrows. Simultaneously, push the tabs into the outer housing so the inner housing begins to slide out. 5 While holding the outer housing, push back against the thumbwheel until the inner housing can be grasped and removed.
Reassemble Adapter Reassembling the adapter is slightly harder because you have to align the connector pins and thumbwheel screw, while sliding the inner housing and outer housing together. 1 Be sure the input connectors and output pins are present and seated on the substrate or PC board.
Disassembling Adapters

Service Removing and Replacing Assemblies
2 Combine the thumbwheel screw and thumbwheel and insert them into the hole in the outer housing. 3 Insert the substrate/board into the outer housing. Slip the attenuator ground (attenuator adapters only) over the thumbwheel screw and seat the input connectors in the proper holes in the housing.

Do not force reassembly of the adapter. The housing halves will slide together with moderate friction. 4 Align the inner housing tabs with the grooves in the side of the outer housing and slide the two partly together. While seating the thumbwheel screw and pins, it will help to hold the assembly vertically, with the input end of the adapter down. This will allow the assemblies to sit vertically and more easily align with the holes in the inner housing. 5 Seat the thumbwheel screw first, then the output pins, into the appropriate holes in the inner housing. 6 Once the two housings are nearly together, press them together firmly until the tabs click into place.
Reassembling the Adapter

Service Removing and Replacing Assemblies
Use the following procedure to disassemble the probe control and power module.
WAR N ING Hazardous voltages exist on the power supply. To avoid electrical shock, adhere closely to the following procedures.
1 Remove the power cord. 2 Remove four flathead screws and remove the top cover.
WAR N ING Be sure to reconnect the safety ground when reassembling the instrument.

This section contains information for ordering parts. Service support for the 1141A Differential Probe is to the assembly level. Service support for the adapters is as complete assemblies, except for the probe tip caps and the probe tips. Service support for the 1142A Probe Control and Power Module is to the component level.
To order a part in the material part list, quote the part number, indicate the quantity desired, and address the order to the nearest Agilent Technologies Sales Office. To order a part not listed in the material list, include the instrument part number, instrument serial number, a description of the part (including its function), and the number of parts required. Address the order to the nearest Agilent Technologies Sales Office.
Service Replaceable Parts

Mail order forms and specific ordering information are available through your local Agilent Technologies Sales Office. Addresses and telephone numbers are located in a separate document shipped with the manuals.
A list of manufacturers’ codes is given the table below. The codes are given for parts in the parts lists. The table gives the manufacturer and address for each code.
Mfr. No. Name Address
00000 Any satisfactory supplier
06665 Precision Monolithics Inc. Santa Clara, CA 95050
24546 Corning Glass Works (Bradford) Bradford, PA 16701
27014 National Semiconductor Corp Palo Alto, CA 94304
28480 Agilent Technologies Corporate Hq Palo Alto, CA 94304
32997 Bourns Inc. Riverside, CA 92507
Service Replaceable Parts


1141A Differential Probe Parts
Service Replaceable Parts


Power and Control Board Component Locator
Service Replaceable Parts


Service Replaceable Parts · Cabling Diagram


IC Connectors not Shown

Supply Pin No. IC Group +15 14 U3 -15 3 GND 15 +15 7 U6-8, 11 -15 4 NC 1, 5, 8 +5 16 U9 GND 8
1142A Probe Control and Power Module


1142A Probe Control and Power Module

Ref. Des.
Qty Description Mfr. Code
Mfr. Part
E1 01141-26102 2 PROBING PIN 28480 01141-26102
E2 01141-22301 1 GROUND BLACK 28480 01141-22301
E3 01141-22401 1 GROUND SCREW 28480 01141-22401
MP1 01141-24102 2 CAP-PROBING PIN 28480 01141-24101
MP2 01141-44702 1 PROBE BOTTOM COVER 28480 01141-44701
MP3 01141-44102 1 PROBE TOP COVER 28480 01141-44101
MP4 01141-22502 1 PROBE POS CLAMP RING 28480 01141-22501
W1 01141-61603 1 PROBE CABLE ASSEMBLY 28480 01141-61601
01141-26102 2 PROBING PIN 28480 01141-26102
01141-24102 2 CAP-PROBING PIN 28480 01141-24101
01141-26102 2 PROBING PIN 28480 01141-26102
01141-24102 2 CAP-PROBING PIN 28480 01141-24101
01141-26102 2 PROBING PIN 28480 01141-26102
01141-24102 2 CAP-PROBING PIN 28480 01141-24101
1142A PROBE CONTROL AND POWER MODULE

Ref. Des.
Qty Description Mfr. Code
Mfr. Part
A1 01142-66501 1 PC ASSEMBLY-POWER AND CONTROL 28480 01142-66501
H1 0515-0374 7 SCREW-MACHINE M3 10mm-LG 00000 ORDER BY DESP.
H2 0515-1031 4 SCREW-MACHINE M3 6mm-LG 90-DEG-FLH-HD 00000 ORDER BY DESP.
H3 0515-1579 1 SCREW-MACHINE M5 18mm-LG 00000 ORDER BY DESP.
MP2 01142-44101 1 TOP CHASSIS 28480 01142-44101
MP3 01142-24701 2 SPACER-HEAT SINK 28480 01142-24701
MP4 0370-1097 2 KNOB-POINTER 28480 0340-1097
MP5 0403-1012 4 FOOT 28480 0403-0727
MP6 5041-0234 5 KEYCAP 28480 5041-0234
W1 8120-1521 1 POWER CORD 18-AWG 3-COND 90-IN-LG (US/ Canada only)

Ref. Des. Part Number
Qty Description Mfr. Code
Mfr. Part
Prefix the reference designators with A1
C1 0160-6190 2 CAPACITOR-FXD 0.33UF ±10% 50VDC 28480 0160-6190
C2 0160-5474 1 CAPACITOR-FXD 0.1UF ±5% 100VDC MET-POLY 28480 0160-5474
C3 0180-3298 2 CAPACITOR-FXD 2200UF+30-10% 50VDC AL 28480 0180-3298
C5 0160-6500 4 CAPACITOR-FXD 0.01UF ±10% 100VDC CER 28480 0160-6500
C6 0160-6500 CAPACITOR-FXD 0.01UF ±10% 100VDC CER 28480 0160-6500
C7 0160-6500 CAPACITOR-FXD 0.01UF ±10% 100VDC CER 28480 0160-6500
C8 0160-6500 CAPACITOR-FXD 0.01UF ±10% 100VDC CER 28480 0160-6500
C9 0180-3845 4 CAPACITOR-FXD 4.7UF ±10% 35VDC TA 28480 0180-3845
C10 0180-3845 CAPACITOR-FXD 4.7UF ±10% 35VDC TA 28480 0180-3845
C12 0160-5581 1 CAPACITOR-FXD 0.033UF ±10% 63VDC 28480 0160-5581
C14 0160-7060 1 CAPACITOR-FXD 3.3UF ±10% 63VDC 28480 0160-7060
C15 0160-4801 1 CAPACITOR-FXD 100PF ±5% 100VDC CER 28480 0160-4801
C16 0180-3845 CAPACITOR-FXD 4.7UF ±10% 35VDC TA 28480 0180-3845
C18 NOT ASSIGNED
C19 0160-5471 1 CAPACITOR-FXD 0.1UF ±5% 50VDC MET-POLY 28480 0160-5471
C20 0160-5469 1 CAPACITOR-FXD 1UF ±10% 50VDC MET-POLY 28480 0160-5469
C23 0180-3784 4 CAPACITOR-FXD 22UF ±20% 25VDC TA 28480 0180-3784
C24 0180-3784 CAPACITOR-FXD 22UF ±20% 25VDC TA 28480 0180-3784
C25 0180-3784 CAPACITOR-FXD 22UF ±20% 25VDC TA 28480 0180-3784
C26 0180-3784 CAPACITOR-FXD 22UF ±20% 25VDC TA 28480 0180-3784
Service Replaceable Parts · Number

Ref. Des. Part Number
Qty Description Mfr. Code
Mfr. Part
CR2 1901-1087 2 DIODE-PWR RECT 100V 3A 04713 MR501
CR4 NOT ASSIGNED
CR5 1901-0731 8 DIODE-PWR RECT 400V 1A 28480 1901-0731
CR6 1901-0731 8 DIODE-PWR RECT 400V 1A 28480 1901-0731
CR9 1901-0731 8 DIODE-PWR RECT 400V 1A 28480 1901-0731
CR10 1901-0731 8 DIODE-PWR RECT 400V 1A 28480 1901-0731
CR11 1901-0731 8 DIODE-PWR RECT 400V 1A 28480 1901-0731
CR12 1901-0731 8 DIODE-PWR RECT 400V 1A 28480 1901-0731
CR13 1901-0734 2 DIODE-PWR RECT 1N5818 30V 1A 04713 IN5818
CR14 1901-0734 2 DIODE-PWR RECT 1N5818 30V 1A 04713 IN5818
CR17 1901-0731 DIODE-PWR RECT 400V 1A 28480 1901-0731
CR18 1901-0731 DIODE-PWR RECT 400V 1A 28480 1901-0731
DS1 1990-0521 1 LED-LAMP LUM-INT=2.2MCD IF=50MA-MAX 24840 5082-4955
E1 2110-0642 1 FUSEHOLDER 6.3A 250A 28480 2110-0642
E2 2110-0565 1 FUSEHOLDER CAP 28480 2110-0565
F1 2110-0201 1 FUSE 0.25A 250V TD 28480 2110-0201
H1 0515-1579 1 SCREW-MACHINE M5 18mm-LG 28480 0515-1579
Service Replaceable Parts · Number

Ref. Des. Part Number
Qty Description Mfr. Code
Mfr. Part
J2 1252-1487 1 CONN-RECT D-SUBMIN 9-CKT (remote) 28480 1252-1487
J3 1252-3935 1 CONNECTOR-ROUND 8-CKT (probe) 28480 1252-3134
MP2 1205-0732 4 SPRING CLIP 28480 1205-0732
MP3 0361-0685 3 RIVET-BLIND DR-PIN RNDH 0.125DIA 28480 0361-0685
MP5-6 NOT ASSIGNED
MP7 1400-0249 1 CABLE TIE 0.062-0.625-DIA 0.091-WD NYL 16956 08-465/GRY
MP9 01142-28801 1 WASHER-TRANSFORMER SUPPORT 28480 01142-28801
R1-2 NOT ASSIGNED
R3 0757-0442 4 RESISTOR 10K 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-1002-F
R4 0757-0465 1 RESISTOR 100K 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-1003-F
R5 0757-0199 5 RESISTOR 21.5K 1% 0.125W TC=0±100 24546 CT4-1/8-TO-2152-F
R6 0757-0199 RESISTOR 21.5K 1% 0.125W TC=0±100 24546 CT4-1/8-TO-2152-F
R7 0698-4431 2 RESISTOR 2.05K 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-2052-F
R8 0699-1203 4 RESISTOR 120.0 1% 0.125W TF TC=0±25 28480 0699-1203
R9 0757-0434 2 RESISTOR 3.65K 1% 0.125W TF TC=±100 24546 CT4-1/8-TO-3651-F
R10 0698-4431 RESISTOR 2.05K 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-2052-F
R11 0699-1203 RESISTOR 120.0 1% 0.125W TF TC=0±25 28480 0699-1203
R12 0757-0434 RESISTOR 3.65K 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-3651-F
R13 0757-0420 1 RESISTOR 750 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-751-F
R14 0698-4002 2 RESISTOR 5K 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-5001-F
R15 0699-1203 RESISTOR 120.0 1% 0.125W TF TC=0±25 28480 0699-1203
Service Replaceable Parts · Number

Ref. Des. Part Number
Qty Description Mfr. Code
Mfr. Part
R16 0698-6317 2 RESISTOR 500 0.1% 0.125W TF TC=0±25 28480 0698-6317
R17 0698-4002 RESISTOR 5K 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-5001-F
R18 0699-1203 RESISTOR 120.0 1% 0.125W TF TC=0±25 28480 0699-1203
R19 0698-6317 RESISTOR 500 0.1% 0.125W TF TC=0±25 28480 0698-6317
R21 0698-8827 1 RESISTOR 1M 1% 0.125W TF TC=0±100 28480 0698-8827
R22 0757-0442 RESISTOR 10K 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-1002-F
R23 0757-0427 1 RESISTOR 1.5K 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-1501-F
R24 0757-0401 1 RESISTOR 100 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-101-F
R25 2100-3161 1 RESISTOR-TRMR 20K 10% TKF SIDE-ADJ 32997 3006P-1-203
R26 2100-3056 1 RESISTOR-TRMR 5K 10% TKF SIDE-ADJ 17-TRN 32997 3006P-1-502
R27 0757-0199 RESISTOR 21.5K 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-2152-F
R28 0757-0199 RESISTOR 21.5K 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-2152-F
R29 0757-0442 RESISTOR 10K 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-1002-F
R30 0757-0442 RESISTOR 10K 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-1002-F
R30 0757-0280 1 RESISTOR 1K 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-1001-F
R32 0757-0460 1 RESISTOR 61.9K 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-6192-F
R40 0757-0123 1 RESISTOR 34.8K 1% 0.125W TF TC=0±100 28480 0757-0123
R42 2100-4250 RESISTOR-VAR 10K 20% 28480 2100-4250
R43 0757-0458 1 RESISTOR 51.1K 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-5112-F
R44 0683-2755 1 RESISTOR 2.7M 1% 0.25W TF TC=0±100 28480 0683-2755
R45 0698-4517 1 RESISTOR 127K 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-1273-F
R46 0698-3271 1 RESISTOR 115K 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-1153-F
R47 0698-8961 1 RESISTOR 909K 1% 0.125W TF TC=0±100 28480 0698-8961
R48 0757-0199 RESISTOR 21.5K 1% 0.125W TF TC=0±100 24546 CT4-1/8-TO-2152-F
Service Replaceable Parts · Number

Ref. Des. Part Number
Qty Description Mfr. Code
Mfr. Part
S1 3101-2609 1 SWITCH-SL DPST STD 5A 250VAC PC 28480 3101-2609
T1 9100-4750 1 TRANSFORMER-POWER (with mtg. hardware) 28480 9100-4750
U1 1826-1403 2 IC V RGLTER-ADJ-POS 3/40V 00000 LT317AT
U2 1826-1670 2 IC V RGLTR-ADJ-NEG -37/1.2V TO-220 PKG 00000 LT337AT
U3 5081-9233 1 ANALOG MULTIPLEXER 4 CHNL 16 -DIP-P 28480 1820-2182
U4 1826-1403 IC V RGLTR ADJ-POS 3/40V 00000 LT317AT
U5 1826-1670 IC V RGLTR-ADJ-NEG -37/1.2V TO-220 PKG 00000 LT337AT
U6 1826-1381 3 IC OP AMP LOW-BIAS-H-IMPD 8-DIP-P PKG 00000 LT1012CN8
U7 1826-1381 IC OP AMP LOW-BIAS-H-IMPD 8-DIP-P PKG 00000 LT1012CN8
U8 1826-1381 IC OP AMP LOW-BIAS-H-IMPD 8-DIP-P PKG 00000 LT1012CN8
U9 1820-3177 1 IC MUXR/DATA-SEL CMOS/74HC 2-TO-1-LINE 04713 MC74HC157N
U10 1826-0774 1 IC V RGLTR-V-REF-FXD 1.22/1.24V TO-92 27014 LM385BZ-1.2
U1 1826-0635 1 IC OP AMP LOW-OFS 8-DIP-P PKG 06665 OP-07CP
VR1 1902-0951 1 DIODE-ZNR 5.1V 5% D0-35 PD=.4W TC±.035% 28480 1902-0951
W1 01141-61602 1 CABLE ASSEMBLY, SAFETY GROUND 28480 01141-61602
Service Replaceable Parts · Number

Ref. Des.
C1 C-2 C17 C-6 CR10 C-5 J1 G-7 R5 B-6 R22 B-3 R45 C-3 U5 D-6
C2 F-3 C18 B-5 CR11 C-3 J2 B-7 R6 B-6 R23 B-5 R46 C-3 U6 B-3
C3 H-2 C19 B-6 CR12 C-6 J3 C-7 R7 C-4 R24 B-5 R47 C-2 U7 C-3
C4 J-3 C20 B-5 CR13 C-3 R8 C-5 R25 B-2 R48 D-3 U8 B-3
C5 C-3 C21 B-5 CR14 C-6 MP1 H-1 R9 C-4 R26 B-2 U9 A-5
C6 C-5 C22 B-6 CR17 B-5 MP2 E-3 R10 C-5 R27 A-5 RP1 B-3 U10 C-2
C7 B-3 C23 B-4 CR18 B-6 MP3 D-3 R11 C-5 R28 A-5 U11 C-3
C8 C-5 C24 B-6 MP4 E-5 R12 C-5 R29 B-6 S1 H-7
C9 C-5 C25 B-4 DS1 G-2 MP7 E-4 R13 E-3 R30 B-6 S2 E-2 VR1 E-3
C10 C-5 C26 G-6 MP8 G-4 R14 C-4 R31 B-3
C11 E-3 E1 F-6 MP9 G-4 R15 C-4 R32 B-2 T1 H-5
C12 A-3 CR2 F-3 E2 E-6 R16 C-4 R40 C-2
C13 A-3 CR3 F-3 P1 E-7 R17 A-2 R41 A-2 U1 D-4
C14 B-4 CR5 C-3 F1 E-6 R18 C-6 R42 C-2 U2 D-5
C15 A-3 CR6 C-5 R3 A-6 R19 C-6 R43 C-2 U3 A-2
C16 C-4 CR9 C-3 HS1 D-5 R4 A-6 R21 B-3 R44 C-3 U4 D-3


Warranty The material contained in this document is provided “as is,” and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control.
commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR 52.227- 14 (June 1987) or DFAR 252.227- 7015 (b)(2) (November 1995), as applicable in any technical data.
© Agilent Technologies, Inc. 2000-2004
No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws.
Manual Part Number 01141-97002, July 2004
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met. WARNING:
01141-97000, June 2000 01141-97001, September 2002 01141-97002, July 2004
Agilent Technologies, Inc. 1900 Garden of the Gods Road Colorado Springs, CO 80907 USA
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
Technology Licenses The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
Restricted Rights Legend If software is for use in the performance of a U.S. Government prime contract or subcontract, Software is delivered and licensed as “Commercial computer software” as defined in DFAR 252.227-7014 (June 1995), or as a “commercial item” as defined in FAR 2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard
Print History

any unintended operation.
Service instructions are for trained service personnel. To avoid dangerous electric shock, do not perform any service unless qualified to do so. Do not attempt internal service or adjustment unless another person, capable of rendering first aid and resuscitation, is present.
This apparatus has been designed and tested in accordance with IEC Publication 1010, Safety Requirements for Measuring Apparatus, and has been supplied in a safe condition. This is a Safety Class I instrument (provided with terminal for protective earthing). Before applying power, verify that the correct safety precautions are taken (see the following warnings). In addition, note the external markings on the instrument that are described under "Safety Symbols."
Instruction manual symbol: the product is marked with this symbol when it is necessary for you to refer to the instruction manual in order to protect against damage to the product.
Do not install substitute parts or perform any unauthorized modification to the instrument.
Capacitors inside the instrument may retain a charge even if the instrument is disconnected from its source of supply.
Hazardous voltage symbol.
Earth terminal symbol: Used to indicate a circuit common connected to grounded chassis.
Do not operate the instrument in the presence of flammable gasses or fumes. Operation of any electrical instrument in such an environment constitutes a definite safety hazard.
Warnings Before turning on the instrument, you must connect the protective earth terminal of the instrument to the protective conductor of the (mains) power cord. The mains plug shall only be inserted in a socket outlet provided with a protective earth contact. You must not negate the protective action by using an extension cord (power cable) without a protective conductor (grounding). Grounding one conductor of a two-conductor outlet is not sufficient protection.
Do not use the instrument in a manner not specified by the manufacturer.
To clean the instrument If the instrument requires cleaning: (1) Remove power from the instrument. (2) Clean the external surfaces of the instrument with a soft cloth dampened with a mixture of mild detergent and water. (3) Make sure that the instrument is completely dry before reconnecting it to a power source.
Only fuses with the required rated current, voltage, and specified type (normal blow, time delay, etc.) should be used. Do not use repaired fuses or short-circuited fuseholders. To do so could cause a shock or fire hazard.
If you energize this instrument by an auto transformer (for voltage reduction or mains isolation), the common terminal must be connected to the earth terminal of the power source.
Whenever it is likely that the ground protection is impaired, you must make the instrument inoperative and secure it against