Air-Oxygen Blender
Model No. PM5200 Series

PM5300 Series (shown)



RECEIVING / INSPECTION Remove the Precision Medical, Inc. Air-Oxygen Blender from the packaging and inspect for damage. If there is any damage, DO NOT USE and contact your Provider. Intended Use Precision Medical, Inc. Air-Oxygen Blender dispenses a continuous and precise blend of medical air and USP oxygen via outlet ports to infant, pediatric and adult patients. The exact Fractional Concentration of Inspired Oxygen (FiO2) blend of gases corresponds to the dialed in FiO2 setting indicated by the control knob (dial). READ ALL INSTRUCTIONS BEFORE USING This manual instructs a Professional to install and operate the Air-Oxygen Blender. This is provided for your safety and to prevent damage to the Air-Oxygen Blender. If you do not understand this manual, DO NOT USE the Air-Oxygen Blender and contact your Provider.
FiO2 Fractional Concentration of Inspired Oxygen
DISS Diameter Indexed Safety System
NIST Non-Interchangeable Screw Thread
psi Pounds Per Square Inch
lpm Liters Per Minute

Symbol for “USE NO OIL”
Symbol indicates the device complies with the requirements of Directive 93/42/EEC concerning medical devices and all applicable International Standards. (On CE marked Devices ONLY)
WARNING Only trained, qualified medical personnel under the direct supervision of a licensed physician should operate the Air-Oxygen Blender . Use this Air-Oxygen Blender only for its Intended Use as described in this manual. Confirm prescribed dose before administering to patient. Monitor on a frequent basis. The Air-Oxygen Blender shall be serviced by a qualified service technician. Always follow ANSI and CGA standards for Medical Gas Products, Flowmeters and Oxygen Handling.

WARNING An Oxygen Analyzer/Monitor must be used to verify oxygen concentration. Accuracy of oxygen concentration will be affected if bleed is not activated at flow settings below 15 lpm for the High Flow Blender, and 3 lpm for the Low Flow Blender. DO NOT obstruct the alarm. DO NOT use Blender when alarm is sounding. DO NOT use oil in or around the Blender. DO NOT occlude or obstruct the bleed port on the auxiliary outlet of the Blender. DO NOT use near any type of flame or flammable/ explosive substances, vapors or atmosphere. Oxygen Concentration Dial does not rotate 360 degrees. Rotating the dial less than 21% or over 100% oxygen will damage the Blender.
CAUTION Turn off gas supplies when Air-Oxygen Blender is not in use. Store the Air-Oxygen Blender in a clean, dry area when not in use. The Air-Oxygen Blender contains magnetic, ferrous material that may affect the results of an MRI. Ensure all connections are tight and leak free. Avoid excessive pressure surges greater than 100 psi (6.9 bar) when pressuring the Blender inlets. DO NOT steam autoclave. DO NOT immerse Air-Oxygen Blender into any liquid. DO NOT gas sterilize with (ETO) Ethylene Trioxide. DO NOT use if dirt or contaminants are present on or around the Blender or connecting devices. DO NOT smoke in an area where oxygen is being administered. DO NOT clean with aromatic hydrocarbons.

Alarm Reset: When pressure differential is 6 psi (0.4 bar) or less. Alarm Sound Level: ≥ to 80 db at 1 ft (0.3 m)
Gas Supply Pressure: 30 - 75 psi (2.1 - 5.2 bar) Air and Oxygen within 10 psi (0.67 bar) of each other Mixed Gas Stability: ±1% Oxygen
Connection Types: DISS Type - Air & Oxygen Inlets & Outlets and / or NIST Type - Air & Oxygen Inlets

Operating Temperature Range: 59°F to 104°F (15°C to 40°C) Transport / Storage Requirements Temperature Range: -10°F to 140°F (-23°C to 60°C)
FiO2 Accuracy:* ± 3% of full scale @ 50 psi (3.4 bar) ± 4% of full scale @ 60 psi (4.14 bar) Pressure Drop:
Low Flow: ≤ 2 psi (0.14 bar) at inlet pressures from 30-90 psi (2.1- 6.2 bar) and at 10 lpm flow rate at 60% FiO2. High Flow: ≤ 3 psi (0.21 bar) at inlet pressures from 30-90 psi (2.1- 6.2 bar) and at 30 lpm flow rate at 60% FiO2.
The Air-Oxygen Blender has been degreased for Oxygen Service prior to delivery.
The Air-Oxygen Blender reverse gas flow complies with clause 6 of ISO 11195.
The Oxygen Analyzer should comply with ISO 7767 to meet CE requirement. Dryness and Composition for inlet gases:
Air: Medical Air supply should meet the requirements of ANSI Z86.1 - 1973 commodity specification for Air, type 1 grade D or better. Oxygen: Oxygen supply must meet all requirements of USP Medical Oxygen Grade N. Dew Point:
Both inlets should remain 10°F (5.5°C) or more below the lowest temperature to which the air distribution system equipment is exposed. At a temperature of 25°F (-3.9°C) and a pressure of 90 psi (6.33 kg/cm2) this equates to 2000 mg/m3.
(ONLY for CE requirements)
* Accuracy of oxygen concentration will be affected if bleed is not activated at flow settings below 15 lpm for the high flow Blender, and 3 lpm for the low flow Blender.
Specifications are subject to change without prior notice.
Specifications continued ·

Depending on model, your fittings may differ from these diagrams.


A
E
F
B G




PM5300 Model · PM5200 Model


Primary Outlets Flow capacity:
This will bleed off some of the air/oxygen mixture to
maintain concentration accuracy at the Low Flow setting.
• · •

Alarm Test 1. The Air-Oxygen Blender is installed and the flowmeter is turned on. Disconnect or turn off the air supply line to the Air-Oxygen Blender. 2. The Blender should alarm with a loud whistle noise. The whistle indicates the alarm is operating correctly. 3. Reconnect and activate the air supply line to the Blender, the alarm should stop whistling. 4. Disconnect or turn off the oxygen supply line to the Blender. 5. The Blender should alarm with a loud whistle noise. The whistle indicates the alarm is operating correctly. 6. Reconnect and activate the oxygen supply line to the Blender, the alarm should stop whistling. 7. If alarm fails to function properly, DO NOT USE. Reverse Gas Flow Procedure (CE Requirements ONLY) 1. Disconnect the oxygen hose from the gas source. Remove all outlet connections from the Blender to ensure that there is no outlet flow. 2. Place the free end of the oxygen supply hose under water. While gradually increasing the air supply pressure from 30-75 psi (2.07- 5.17 bar) check for leakage past the oxygen inlet check valve. 3. Replace the Duckbill Check Valve in the oxygen inlet if bubbles indicate leakage. Reference Air-Oxygen Blender Service Manual (P/N 504827.) 4. Repeat steps 1-3 to check for leakage past the air inlet check valve. OPERATING INSTRUCTIONS
360°. DO NOT force dial less than 21% or over 100% oxygen, this will damage the Blender.

3. To activate the bleed, turn the knurled collar until the bleed pin is engaged. Then slide the collar back until it contacts the cover. 4. Confirm the concentration of air/oxygen with an Oxygen Analyzer/Monitor. 5. Confirm the flow of air and/or oxygen mixture to the patient. 6. Turn “OFF” the Air and Oxygen supply when deactivating the Auxiliary bleed, and no device is attached to the outlet port. 7. To deactivate the bleed, pull the collar out away from the cover while rotating the collar until the slot engages the pin. Then slide the collar past the pin and turn an additional 1/4 turn to lock in place. NOTE: This Procedure is completed with ease when all gas
supplies to the Blender are turned “OFF”. 8. Turn “OFF” the Air and Oxygen supply or disconnect when the Blender is not in use. 9. Operational Verification listed below should be done before the Blender is placed in service.
Verification consists of:
Performance Check A detailed description of the tests listed below can be found in the
Blender Service Manual (P/N 504827), available on the Internet; www.precisionmedical.com

Every month verify proper functioning of alarm. Every year conduct the Operational Verification Procedure (OVP). Every 2 years the Air-Oxygen Blender should be serviced. PM5200 (P/N 505407) PM5300 (P/N 504932) Refer to the Air-Oxygen Blender Service Manual (P/N 504827) for complete details regarding further maintenance and testing.
Technical Description For a complete Technical Description of the Air-Oxygen Blender and list of Replacement Parts, reference the Air-Oxygen Blender Service Manual (P/N 504827) available on the Internet; www.precisionmedical.com Returns Returned products require a Returned Goods Authorization (RGA) number, contact Precision Medical, Inc. All returns must be packaged in sealed containers to prevent damage. Precision Medical, Inc. will not be responsible for goods damaged in transit. Refer to Precision Medical, Inc. Return Policy available on the Internet; www.precisionmedical.com Disposal Instructions This device and its packaging contain no hazardous materials. No special precautions need to be taken when disposing the device and/or its packaging.
Please Recycle

Troubleshooting If the Air-Oxygen Blender fails to function, consult the Troubleshooting Guide below. If problem cannot be solved by using Troubleshooting Guide refer to the Air-Oxygen Blender Service Manual (P/N 504827) available on the Internet; www.precisionmedical.com or consult your Provider.
Problem Probable Cause Remedy
Oxygen concentration discrepancy between Blender setting and Analyzer/Monitor (greater than 3%)
below 3 lpm. 2. Analyzer/Monitor inaccurate
No flow at Blender outlets
Alarm sounding 1. Difference between Oxygen and air inlet pressures greater than specified

AND LIMITATION OF LIABILITY Precision Medical, Inc. warrants that the Blender, (the Product), will be free of defects in workmanship and/or material for the following period:
Two (2) years from shipment
Should any failure to conform to this warranty appear within the applicable period, Precision Medical, Inc. shall, upon written notification thereof and substantiation that the goods have been stored, installed, maintained and operated in accordance with Precision Medical, Inc.’s instructions and standard industry practice, and that no modifications, substitutions, or alterations have been made to the goods, correct such defect by suitable repair or replacement at its own expense.
ORAL STATEMENTS DO NOT CONSTITUTE WARRANTIES.
The representatives of Precision Medical, Inc. or any retailers are not authorized to make oral warranties about the merchandise described in this contract, and any such statements shall not be relied upon and are not part of the contract for sale. Thus, this writing is a final, complete and exclusive statement of the terms of that contract.
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ANY WARRANTY OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE OR OTHER WARRANTY OF QUALITY, WHETHER EXPRESS OR IMPLIED.
Precision Medical, Inc. shall not under any circumstances be liable for special, incidental or consequential damages including but not limited to lost profits, lost sales, or injury to person or property. Correction of non- conformities as provided above shall constitute fulfillment of all liabilities of Precision Medical, Inc. whether based on contract, negligence, strict tort or otherwise. Precision Medical, Inc. reserves the right to discontinue manufacture of any product or change product materials, designs, or specifications without notice.
Precision Medical, Inc. reserves the right to correct clerical or typographical errors without penalty.

Molenstraat 15 2513 BH, The Hague The Netherlands
As delivered, the object of the declaration described above is in conformity with the requirements of MDD 93/42/EEC Annex II.3 and the following documents:
ISO 14971 Medical Devices - Application of Risk Management 2000+A1:2003 to Medical Devices
EN 980 Graphical Symbols for Use in the Labeling of Medical 2003 Devices
EN 1041 Information supplied by the Manufacturer with Medical 1998 Devices
ISO 11195 Gas mixers for medical use - stand alone gas mixers 1995
ISO 15001 Anesthetic and respiratory equipment -- Compatibility 2004 with oxygen
EC Certificate No.: HD600119110 0001
504745 Rev8 (E) 03/18/08 (?M) Printed in USA
Document Title Edition · ISO 13485 Certified
Common problems, likely causes, and what to do.
| Problem | Possible cause | Remedy | |
|---|---|---|---|
| Oxygen concentration discrepancy between Blender setting and Analyzer/Monitor (greater than 3%) | 1. High Flow model, flow requirement below 15 lpm. Low Flow model, flow requirement below 3 lpm. 2. Analyzer/Monitor inaccurate 3. Low flow bleed obstructed 4. Gas supply contaminated 5. Downstream device causing back flow or restricted flow | 1. Use auxiliary outlet & engage bleed 2. Recalibrate Analyzer/Monitor or Verify with second Analyzer/Monitor 3. Remove obstruction 4. Check gas sources with calibrated Oxygen Analyzer/ Monitor to confirm Oxygen is 100% and Air is 21% 5. Isolate Blender. Check oxygen concentration at Blender Outlets | p.14 |
| No flow at Blender outlets | 1. Gas sources turned “OFF” 2. Gas sources not connected | 1. Turn gas sources “ON” 2. Connect gas sources | p.14 |
| Alarm sounding | 1. Difference between Oxygen and air inlet pressures greater than specified | 1. Correct pressure difference until Air and Oxygen pressures are within specification | p.14 |