U S E R M A N U A L

Air-Oxygen Blender

(DISS and NIST Connections)

Model No. PM5200 Series

Precision Trading PM5200 — (DISS and NIST Connections) diagram (page 1)
Fig. 1 · (DISS and NIST Connections)

SAVE THESE INSTRUCTIONS

300 Held Drive Tel(+001) 610-262-6090 Northampton, PA 18067 USA Fax: (+001) 610-262-6080 www.precisionmedical.com

PM5300 Series (shown)

Page 2 of the Precision Trading PM5200 user manual — U S E R M A N U A L
Page 2 · click to enlarge
Page 2
Page 2 of the Precision Trading PM5200 user manual — U S E R M A N U A L
Page 3 of the Precision Trading PM5200 user manual — Explanation of Abbreviations
Page 3 · click to enlarge
Page 3

RECEIVING / INSPECTION Remove the Precision Medical, Inc. Air-Oxygen Blender from the packaging and inspect for damage. If there is any damage, DO NOT USE and contact your Provider. Intended Use Precision Medical, Inc. Air-Oxygen Blender dispenses a continuous and precise blend of medical air and USP oxygen via outlet ports to infant, pediatric and adult patients. The exact Fractional Concentration of Inspired Oxygen (FiO2) blend of gases corresponds to the dialed in FiO2 setting indicated by the control knob (dial). READ ALL INSTRUCTIONS BEFORE USING This manual instructs a Professional to install and operate the Air-Oxygen Blender. This is provided for your safety and to prevent damage to the Air-Oxygen Blender. If you do not understand this manual, DO NOT USE the Air-Oxygen Blender and contact your Provider.

Explanation of Abbreviations

FiO2 Fractional Concentration of Inspired Oxygen

DISS Diameter Indexed Safety System

NIST Non-Interchangeable Screw Thread

psi Pounds Per Square Inch

lpm Liters Per Minute

Page 4 of the Precision Trading PM5200 user manual — CONSULT ACCOMPANYING DOCUMENTS
Page 4 · click to enlarge
Page 4

SAFETY Information - Warnings and Cautions

CONSULT ACCOMPANYING DOCUMENTS

Symbol for “USE NO OIL”

Symbol indicates the device complies with the requirements of Directive 93/42/EEC concerning medical devices and all applicable International Standards. (On CE marked Devices ONLY)

WARNING Only trained, qualified medical personnel under the direct supervision of a licensed physician should operate the Air-Oxygen Blender . Use this Air-Oxygen Blender only for its Intended Use as described in this manual. Confirm prescribed dose before administering to patient. Monitor on a frequent basis. The Air-Oxygen Blender shall be serviced by a qualified service technician. Always follow ANSI and CGA standards for Medical Gas Products, Flowmeters and Oxygen Handling.

Show 5 more labels / short text on this page
• · • · • · • · •
Page 5 of the Precision Trading PM5200 user manual — SAFETY Information - Warnings and Cautions
Page 5 · click to enlarge
Page 5

WARNING An Oxygen Analyzer/Monitor must be used to verify oxygen concentration. Accuracy of oxygen concentration will be affected if bleed is not activated at flow settings below 15 lpm for the High Flow Blender, and 3 lpm for the Low Flow Blender. DO NOT obstruct the alarm. DO NOT use Blender when alarm is sounding. DO NOT use oil in or around the Blender. DO NOT occlude or obstruct the bleed port on the auxiliary outlet of the Blender. DO NOT use near any type of flame or flammable/ explosive substances, vapors or atmosphere. Oxygen Concentration Dial does not rotate 360 degrees. Rotating the dial less than 21% or over 100% oxygen will damage the Blender.

CAUTION Turn off gas supplies when Air-Oxygen Blender is not in use. Store the Air-Oxygen Blender in a clean, dry area when not in use. The Air-Oxygen Blender contains magnetic, ferrous material that may affect the results of an MRI. Ensure all connections are tight and leak free. Avoid excessive pressure surges greater than 100 psi (6.9 bar) when pressuring the Blender inlets. DO NOT steam autoclave. DO NOT immerse Air-Oxygen Blender into any liquid. DO NOT gas sterilize with (ETO) Ethylene Trioxide. DO NOT use if dirt or contaminants are present on or around the Blender or connecting devices. DO NOT smoke in an area where oxygen is being administered. DO NOT clean with aromatic hydrocarbons.

Show 10 more labels / short text on this page
• · • · • · • · • • · • · • • · • · • ·
Page 6 of the Precision Trading PM5200 user manual — Specifications
Page 6 · click to enlarge
Page 6

Specifications

ModelPM5200 High FlowPM5300 Low Flow Primary Outlet Flow Range15 - 120 lpm3 - 30 lpm With both supply pressures at 50 psi (3.4 bar) with NO BLEED Auxiliary Outlet Flow Range2 - 100 lpm0 - 30 lpm With both supply pressures at 50 psi (3.4 bar) with BLEED Bleed Flow13 lpm or less at 50 psi (3.4 bar)3 lpm or less at 50 psi (3.4 bar) Maximum Combined Flow (All Outlets)≥ 120 lpm≥ 30 lpm Bypass Flow (Loss of Air or Oxygen supply)> 85 lpm> 45 lpm Bypass Alarm Activation50 psi (3.45 bar)60 psi (4.14 bar)50 psi (3.45 bar)60 psi (4.14 bar) 13-25 psi16-24 psi18-22 psi16-24 psi 0.9-1.7 bar1.1-1.65 bar1.2-1.5 bar1.1-1.65 bar

Alarm Reset: When pressure differential is 6 psi (0.4 bar) or less. Alarm Sound Level: ≥ to 80 db at 1 ft (0.3 m)

Oxygen Concentration Adjustment Range21 - 100%

Gas Supply Pressure: 30 - 75 psi (2.1 - 5.2 bar) Air and Oxygen within 10 psi (0.67 bar) of each other Mixed Gas Stability: ±1% Oxygen

Connection Types: DISS Type - Air & Oxygen Inlets & Outlets and / or NIST Type - Air & Oxygen Inlets

Page 7 of the Precision Trading PM5200 user manual — Specifications
Page 7 · click to enlarge
Page 7
Dimensions(without fittings)
Depth4.9 in (12.5 cm)
Width2.3 in (5.7 cm)
Height4.1 in (10.4 cm)
Weight2.9 lbs (1.3 kg)
Shipping Weight3.5 lbs (1.6 kg)

Operating Temperature Range: 59°F to 104°F (15°C to 40°C) Transport / Storage Requirements Temperature Range: -10°F to 140°F (-23°C to 60°C)

HumidityMax 95% Noncondensing

FiO2 Accuracy:* ± 3% of full scale @ 50 psi (3.4 bar) ± 4% of full scale @ 60 psi (4.14 bar) Pressure Drop:

Low Flow: ≤ 2 psi (0.14 bar) at inlet pressures from 30-90 psi (2.1- 6.2 bar) and at 10 lpm flow rate at 60% FiO2. High Flow: ≤ 3 psi (0.21 bar) at inlet pressures from 30-90 psi (2.1- 6.2 bar) and at 30 lpm flow rate at 60% FiO2.

The Air-Oxygen Blender has been degreased for Oxygen Service prior to delivery.

The Air-Oxygen Blender reverse gas flow complies with clause 6 of ISO 11195.

The Oxygen Analyzer should comply with ISO 7767 to meet CE requirement. Dryness and Composition for inlet gases:

Air: Medical Air supply should meet the requirements of ANSI Z86.1 - 1973 commodity specification for Air, type 1 grade D or better. Oxygen: Oxygen supply must meet all requirements of USP Medical Oxygen Grade N. Dew Point:

Both inlets should remain 10°F (5.5°C) or more below the lowest temperature to which the air distribution system equipment is exposed. At a temperature of 25°F (-3.9°C) and a pressure of 90 psi (6.33 kg/cm2) this equates to 2000 mg/m3.

(ONLY for CE requirements)

* Accuracy of oxygen concentration will be affected if bleed is not activated at flow settings below 15 lpm for the high flow Blender, and 3 lpm for the low flow Blender.

Specifications are subject to change without prior notice.

Specifications continued ·

Page 8 of the Precision Trading PM5200 user manual — H
Page 8 · click to enlarge
Page 8

Diagrams

Depending on model, your fittings may differ from these diagrams.

Precision Trading PM5200 — Diagrams diagram (page 8)
Fig. 2 · Diagrams
Precision Trading PM5200 — Diagrams diagram (page 8)
Fig. 3 · Diagrams

B

A

E

F

B G

Precision Trading PM5200 — B diagram (page 8)
Fig. 4
Precision Trading PM5200 — B diagram (page 8)
Fig. 5

D C

Precision Trading PM5200 — D C diagram (page 8)
Fig. 6
Precision Trading PM5200 — D C diagram (page 8)
Fig. 7

H

PM5300 Model · PM5200 Model

Page 9 of the Precision Trading PM5200 user manual — COMPONENT Description
Page 9 · click to enlarge
Page 9

COMPONENT Description

ITEMDESCRIPTION AOxygen Concentration Dial A dial used for selecting oxygen concentrations between 21%-100%. The FiO scale is used for reference only. 2 This Dial does not rotate 360°. The dial starts at 21% and ends at 100%. BPrimary Outlet Port A male DISS oxygen ftiting with check valve that delivers flow when engaged to any controlling device, such as a flowmeter. CAuxiliary Outlet Port A male DISS oxygen ftiting with check valve that delivers fol w when engaged to any controlling device, such as a fol wmeter. This outlet is equipped with a bleed valve that allows the user to control if the bleed is ON or OFF. With the bleed in the ON position, this outlet delivers accurate oxygen concentrations in the following flows: Model Flow Range High Flow 2 – 100 lpm Low Flow 0 – 30 lpm DAuxiliary Bleed Collar The collar is used to engage and disengage the bleed. The bleed is necessary to maintain accurate FiO 2 Concentration below 15 lpm for the High Flow and 3 lpm for the Low Flow. This collar is designed to prevent accidental disengagement of the bleed. To activate the bleed, turn the knurled collar until the bleed pin is engaged. Slide collar back until it contacts the cover. To deactivate the bleed, pull collar away from cover while rotating until the collar slot engages the pin, turn additional 1/4 turn to lock in place. EOxygen Inlet Fitting A female DISS or NIST oxygen ftiting with one way valve that is used to connect an oxygen supply hose.

Page 10 of the Precision Trading PM5200 user manual — Installation
Page 10 · click to enlarge
Page 10

Component Description

ITEMDESCRIPTION FAir Inlet Fitting A male DISS or NIST air fitting with one way valve that is used to connect an air supply hose. GAlarm An audible alarm that sounds due to an excessive pressure drop or deletion of either gas supply. HManifold Outlet (Optional) Manifold with 3 primary outlets. IRear Slide Mount with dove tail.

Installation

1. Secure the Air-Oxygen Blender to a wall or pole bracket in an upright position. 2. It is recommended to install a condensation trap in the air supply line. 3. Connect the air and oxygen supply lines to the appropriate inlet fittings on the bottom of the Blender. 4. Attach a flowmeter, or other metering device to one of the outlet ports.

Primary Outlets Flow capacity:

  • High Flow Blender (PM 5200 Model) 15 lpm to 120 lpm
  • Low Flow Blender (PM 5300 Model) 3 lpm to 30 lpm Auxiliary Outlet:

This will bleed off some of the air/oxygen mixture to

maintain concentration accuracy at the Low Flow setting.

  • High Flow Blender (PM 5200 Model) 15 lpm or less
  • Low Flow Blender (PM 5300 Model) 3 lpm or less 5. Attach a supply line to the outlet port of the flowmeter.

• · •

Page 11 of the Precision Trading PM5200 user manual — Installation
Page 11 · click to enlarge
Page 11

Alarm Test 1. The Air-Oxygen Blender is installed and the flowmeter is turned on. Disconnect or turn off the air supply line to the Air-Oxygen Blender. 2. The Blender should alarm with a loud whistle noise. The whistle indicates the alarm is operating correctly. 3. Reconnect and activate the air supply line to the Blender, the alarm should stop whistling. 4. Disconnect or turn off the oxygen supply line to the Blender. 5. The Blender should alarm with a loud whistle noise. The whistle indicates the alarm is operating correctly. 6. Reconnect and activate the oxygen supply line to the Blender, the alarm should stop whistling. 7. If alarm fails to function properly, DO NOT USE. Reverse Gas Flow Procedure (CE Requirements ONLY) 1. Disconnect the oxygen hose from the gas source. Remove all outlet connections from the Blender to ensure that there is no outlet flow. 2. Place the free end of the oxygen supply hose under water. While gradually increasing the air supply pressure from 30-75 psi (2.07- 5.17 bar) check for leakage past the oxygen inlet check valve. 3. Replace the Duckbill Check Valve in the oxygen inlet if bubbles indicate leakage. Reference Air-Oxygen Blender Service Manual (P/N 504827.) 4. Repeat steps 1-3 to check for leakage past the air inlet check valve. OPERATING INSTRUCTIONS

1. Turn “ON” the Air and Oxygen supply. 2. Adjust the Oxygen Concentration Dial to the prescribed concentration. NOTE: The Oxygen Concentration Dial does not rotate

360°. DO NOT force dial less than 21% or over 100% oxygen, this will damage the Blender.

Page 12 of the Precision Trading PM5200 user manual — CLEANING
Page 12 · click to enlarge
Page 12

3. To activate the bleed, turn the knurled collar until the bleed pin is engaged. Then slide the collar back until it contacts the cover. 4. Confirm the concentration of air/oxygen with an Oxygen Analyzer/Monitor. 5. Confirm the flow of air and/or oxygen mixture to the patient. 6. Turn “OFF” the Air and Oxygen supply when deactivating the Auxiliary bleed, and no device is attached to the outlet port. 7. To deactivate the bleed, pull the collar out away from the cover while rotating the collar until the slot engages the pin. Then slide the collar past the pin and turn an additional 1/4 turn to lock in place. NOTE: This Procedure is completed with ease when all gas

supplies to the Blender are turned “OFF”. 8. Turn “OFF” the Air and Oxygen supply or disconnect when the Blender is not in use. 9. Operational Verification listed below should be done before the Blender is placed in service.

Verification consists of:

  • Alarm Test
  • Verify FiO 2 Concentration with a Oxygen Analyzer/Monitor
  • Reverse Gas Flow Procedure

Performance Check A detailed description of the tests listed below can be found in the

Blender Service Manual (P/N 504827), available on the Internet; www.precisionmedical.com

  • Reverse Gas Flow Procedure
  • Operational Verification Procedure (OVP)

CLEANING

Page 13 of the Precision Trading PM5200 user manual — CLEANING
Page 13 · click to enlarge
Page 13
1. Disconnect all gas connections and equipment before cleaning. 2. Clean exterior surfaces with a cloth dampened with mild detergent and water. 3. Wipe dry with a clean cloth. Maintenance The following maintenance on the Air-Oxygen Blender must be performed by a trained service technician:

Every month verify proper functioning of alarm. Every year conduct the Operational Verification Procedure (OVP). Every 2 years the Air-Oxygen Blender should be serviced. PM5200 (P/N 505407) PM5300 (P/N 504932) Refer to the Air-Oxygen Blender Service Manual (P/N 504827) for complete details regarding further maintenance and testing.

Technical Description For a complete Technical Description of the Air-Oxygen Blender and list of Replacement Parts, reference the Air-Oxygen Blender Service Manual (P/N 504827) available on the Internet; www.precisionmedical.com Returns Returned products require a Returned Goods Authorization (RGA) number, contact Precision Medical, Inc. All returns must be packaged in sealed containers to prevent damage. Precision Medical, Inc. will not be responsible for goods damaged in transit. Refer to Precision Medical, Inc. Return Policy available on the Internet; www.precisionmedical.com Disposal Instructions This device and its packaging contain no hazardous materials. No special precautions need to be taken when disposing the device and/or its packaging.

Please Recycle

Page 14 of the Precision Trading PM5200 user manual — CLEANING
Page 14 · click to enlarge
Page 14

Troubleshooting If the Air-Oxygen Blender fails to function, consult the Troubleshooting Guide below. If problem cannot be solved by using Troubleshooting Guide refer to the Air-Oxygen Blender Service Manual (P/N 504827) available on the Internet; www.precisionmedical.com or consult your Provider.

Problem Probable Cause Remedy

Oxygen concentration discrepancy between Blender setting and Analyzer/Monitor (greater than 3%)

1. High Flow model, flow requirement below 15 lpm. Low Flow model, flow requirement
1. Use auxiliary outlet & engage bleed

below 3 lpm. 2. Analyzer/Monitor inaccurate

2. Recalibrate Analyzer/Monitor or Verify with second Analyzer/Monitor 3. Remove obstruction
3. Low flow bleed obstructed 4. Gas supply contaminated
4. Check gas sources with calibrated Oxygen Analyzer/ Monitor to confirm Oxygen is 100% and Air is 21% 5. Isolate Blender. Check oxygen concentration at Blender Outlets
5. Downstream device causing back flow or restricted flow

No flow at Blender outlets

1. Gas sources turned “OFF” 2. Gas sources not connected
1. Turn gas sources “ON” 2. Connect gas sources

Alarm sounding 1. Difference between Oxygen and air inlet pressures greater than specified

1. Correct pressure difference until Air and Oxygen pressures are within specification
Page 15 of the Precision Trading PM5200 user manual — LIMITED WARRANTY
Page 15 · click to enlarge
Page 15

LIMITED WARRANTY

AND LIMITATION OF LIABILITY Precision Medical, Inc. warrants that the Blender, (the Product), will be free of defects in workmanship and/or material for the following period:

Two (2) years from shipment

Should any failure to conform to this warranty appear within the applicable period, Precision Medical, Inc. shall, upon written notification thereof and substantiation that the goods have been stored, installed, maintained and operated in accordance with Precision Medical, Inc.’s instructions and standard industry practice, and that no modifications, substitutions, or alterations have been made to the goods, correct such defect by suitable repair or replacement at its own expense.

ORAL STATEMENTS DO NOT CONSTITUTE WARRANTIES.

The representatives of Precision Medical, Inc. or any retailers are not authorized to make oral warranties about the merchandise described in this contract, and any such statements shall not be relied upon and are not part of the contract for sale. Thus, this writing is a final, complete and exclusive statement of the terms of that contract.

THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ANY WARRANTY OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE OR OTHER WARRANTY OF QUALITY, WHETHER EXPRESS OR IMPLIED.

Precision Medical, Inc. shall not under any circumstances be liable for special, incidental or consequential damages including but not limited to lost profits, lost sales, or injury to person or property. Correction of non- conformities as provided above shall constitute fulfillment of all liabilities of Precision Medical, Inc. whether based on contract, negligence, strict tort or otherwise. Precision Medical, Inc. reserves the right to discontinue manufacture of any product or change product materials, designs, or specifications without notice.

Precision Medical, Inc. reserves the right to correct clerical or typographical errors without penalty.

Page 16 of the Precision Trading PM5200 user manual — LIMITED WARRANTY
Page 16 · click to enlarge
Page 16
DECLARATION OF CONFORMITY ManufacturerPrecision Medical, Inc. 300 Held Drive, Northampton, PA 18067, USA CONTACT: Quality Manager
Phone610-262-6090
Authorized European RepresentativeEmergo Europe

Molenstraat 15 2513 BH, The Hague The Netherlands

ProductGas Mixers for Medical Use
Model(s)PM5200EN, PM5200MEN, PM5300EN, PM5300MEN
MDD ClassIIb
Classification criteriaClause 3.1 Rule 11 of Annex IX of MDD

As delivered, the object of the declaration described above is in conformity with the requirements of MDD 93/42/EEC Annex II.3 and the following documents:

ISO 14971 Medical Devices - Application of Risk Management 2000+A1:2003 to Medical Devices

EN 980 Graphical Symbols for Use in the Labeling of Medical 2003 Devices

EN 1041 Information supplied by the Manufacturer with Medical 1998 Devices

ISO 11195 Gas mixers for medical use - stand alone gas mixers 1995

ISO 15001 Anesthetic and respiratory equipment -- Compatibility 2004 with oxygen

Notified BodyTÜV Rheinland Products Safety GmbH

EC Certificate No.: HD600119110 0001

www.precisionmedical.com

504745 Rev8 (E) 03/18/08 (?M) Printed in USA

Document Title Edition · ISO 13485 Certified

Fix-it

Troubleshooting

Common problems, likely causes, and what to do.

ProblemPossible causeRemedy
Oxygen concentration discrepancy between Blender setting and Analyzer/Monitor (greater than 3%)1. High Flow model, flow requirement below 15 lpm. Low Flow model, flow requirement below 3 lpm. 2. Analyzer/Monitor inaccurate 3. Low flow bleed obstructed 4. Gas supply contaminated 5. Downstream device causing back flow or restricted flow1. Use auxiliary outlet & engage bleed 2. Recalibrate Analyzer/Monitor or Verify with second Analyzer/Monitor 3. Remove obstruction 4. Check gas sources with calibrated Oxygen Analyzer/ Monitor to confirm Oxygen is 100% and Air is 21% 5. Isolate Blender. Check oxygen concentration at Blender Outlets p.14
No flow at Blender outlets1. Gas sources turned “OFF” 2. Gas sources not connected1. Turn gas sources “ON” 2. Connect gas sources p.14
Alarm sounding1. Difference between Oxygen and air inlet pressures greater than specified1. Correct pressure difference until Air and Oxygen pressures are within specification p.14

Ask about the Precision Trading PM5200

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the Precision Trading PM5200

Be the first to ask — other owners or our team will answer.

Original document

Precision Trading PM5200 manual PDF

16 pages·358 KB·Respiratory Product·PDF
Click any page above to open it here
Page 1 / …
100%
Loading pages…