Operation Manual
Make sure you read this manual before using the instrument. Keep this manual in a safe place so that you can use it in the future.


Thank you for purchasing the Canon Digital Retinal Camera CR-2 (referred to as “CR-2” in this manual). The operation manuals for this instrument consist of the “Digital Retinal Camera CR-2 Operation Manual” and the “Retinal imaging control software NM 2 Operation Manual.” Be sure to read these manuals thoroughly before using the instrument, and apply the information that you learn.



The Canon Digital Retinal Camera CR-2 takes digital photographic retinal images of patient’s eyes across an angle of view of 45 degrees, in a naturally dilated state without using mydriatics. The CR-2 has the following features:
The CR-2 can take images at ISO 800/1600 at the standard flash intensity, and ISO 3200/6400 in low flash intensity mode. This reduces patient discomfort caused by light glare.
The CR-2 is designed to be even smaller, more lightweight and compact than the previous model. This makes it easier to assist patients and provides significant ease of operation. An operation lever has the integrated functions: vertical movement of the main unit, switching between anterior segment and fundus, and manual focusing. This lever significantly improves alignment operability.
The CR-2 is equipped with a special digital camera using EOS digital technology. This enables the CR-2 to fully maximize Canon technology for shooting high-quality, high-resolution diagnostic images.
The CR-2 captures retinal images through pupils of the normal diameters from 4.0 mm. This small pupil photography function permits image to be captured through pupils from 3.3 mm in diameters.
With the supplied imaging control software, the photographed images can be browsed, processed, stored, output to the DICOM storage server, exported in DICOM or JPEG format and output to a printer. In this manual, the operation procedure is described with the Retinal imaging control software NM 2 (referred to as “Rics NM 2” ).
The device is intended to be used for taking digital images of the retina of the human eye without a mydriatic.
This medical device is intended to observe and record images of retinal fundus through the pupil without making contact with subject’s eye for the purpose of diagnosis by way of producing fundus image information.

External eye fixation lamp Chin rest paper—500 sheets
1 Introduction

Symbols to indicate safety precautions
This manual uses the following symbols to indicate safety precautions that are important for using the CR-2 safely. Always follow the advice in these safety precautions.
This symbol indicates actions that must not be taken (prohibited actions).
This symbol indicates actions that must be taken.
This symbol indicates important advice that we strongly recommend be followed while operating the CR-2.
This symbol indicates supplementary explanations or advice for operating the CR-2.
This manual uses the following style to indicate reference destination. Example: (see page 15)
1 Introduction

The following sections list the classifications applicable to the CR-2 and the directives and standards that the CR-2 complies with.
Protection against electric shock Class I equipment
Degree of protection against electric shock Type B applied parts
Degree of protection against ingress of water IPX0
Degree of safety of application in the presence of a flammable anaesthetic mixture with air or with oxygen or nitrous oxide
Mode of operation Continuous operation
UL 60601-1 CAN/CSA C22.2 No.601.1 IEC 60601-1: 1988+A1: 1991+A2: 1995 IEC 60601-1: 2005
Medical electrical equipment – Part 1: General requirements for safety
IEC 60601-1-1: 2000 Medical electrical equipment – Part 1-1: General requirements for safety – Collateral standard: Safety requirements for medical electrical systems
IEC60601-1-2: 2001+A1: 2004 Medical electrical equipment – Part 1-2: Collateral standard: Electromagnetic compatibility – Requirements and tests
Biological evaluation of medical devices – Evaluation and testing
Not suitable

93/42/EEC Medical Device Directive
Medical electrical equipment – Part 1: General requirements for safety
IEC 60601-1-1: 2001 Medical electrical equipment – Part 1-1: General requirements for safety – Collateral standard: Safety requirements for medical electrical systems
EN 60601-1-2: 2007 Medical electrical equipment – Part 1-2: Collateral standard: Electromagnetic compatibility – Requirements and tests
IEC60601-1-4: 1996+A1: 1999 Medical electrical equipment – Part 1-4: General requirements for safety – Collateral standard: Programmable electrical medical systems
EN 60601-1-6: 2004 Medical electrical equipment – Part 1-6: General requirements for safety – Collateral standard: Usability
Biological evaluation of medical devices – Evaluation and testing
Während des Betriebs liegt der Schalldruckpegel dieses Instruments bei 70 dB(A) oder weniger gemäß EN ISO 7779.
2 Safety

To prevent injuries and data loss, operate the CR-2 correctly by following the safety precautions.
Do not damage the power cord.
Handle the power cord carefully. If the cord is damaged, contact your sales representative or local Canon dealer. A damaged cord may result in fire or electric shock.
Do not use a multiple power strip and an extension cable.
Connect the power cord directly to the AC outlet. Do not use a multiple power strip and an extension cable with it.
Do not disassemble or modify.
A disassembled or modified instrument may result in fire or electric shock. Because the CR-2 incorporates high-voltage parts that may cause electric shocks, touching them may cause death or serious injury.
Do not leave alcohol, thinner, or any flammable chemicals near the instrument.
Do not place the CR-2 near a flammable solvent. Fire may result if the solvent spills or evaporates and makes contact with the internal electric parts. Some disinfectants are flammable. Be sure to take precautions when using them.
Stop using immediately if there is an abnormality or problem. If an abnormality occurs, immediately turn off the power and disconnect the plugs on all devices.
Fire or electric shock may result if you continue using in such conditions. Immediately turn off the power of CR-2, unplug the power plug and turn off the power of all connected devices. Then, contact your sales representative or local Canon dealer immediately.
Do not place anything on top of the device.
Fire or electric shock may result if water or any other liquid or a needle, paper clip or any foreign matter gets inside the CR-2.

Do not use a power supply voltage other than that specified on the rating label. Do not use any power cord other than the one supplied.
Use the power voltage specified on the rating label. Otherwise, fire or electric shock may result. The supplied power cord is designed exclusively for this product. Do not use any other power cord.
Do not plug or unplug the power plug with wet hands.
Do not plug or unplug the power plug or handle any other parts with wet hands. Otherwise, fire or electric shock may result.
Unplug the plug periodically and remove any dust or dirt around the plug and the AC outlet.
If the cord is kept plugged in for a long time in a dusty, humid or sooty environment, dust around the plug will attract moisture, resulting in possible insulation failure that may result in a fire.
Insert the power plug completely.
Insert the power plug completely into the AC outlet. If a pin of the power plug makes contact with metal or any conductive object, fire or electric shock may result.
Do not clean the CR-2 with flammable solvent.
For safety reasons, before cleaning the CR-2, be sure to turn off the power of all the devices and unplug the power cord from the AC outlet. Do not use alcohol, benzine, thinner or any other flammable solvent. Otherwise, fire or electric shock may result.
Turn off the power before inspection.
For safety reasons, before inspecting the instrument or cables, be sure to turn off the power of all the devices. Otherwise, electric shock may result.
Do not play the supplied CD-ROM in any CD player that does not support data CD-ROMs.
Playing the CD-ROM in an audio CD player may damage the speaker. Playing the CD-ROM in an audio CD player may also cause hearing loss if listened to with headphones.
Be sure to turn off the power before moving the CR-2.
Before moving the CR-2, make sure that the power is turned off, the power plug is removed from the AC outlet, and any cables that are connected to other devices are disconnected.
Do not touch conductive parts of non-medical equipment and the patient simultaneously.
Otherwise, electric shock may result.
Use an isolation transformer when this system is connected to a network.
Connect an isolation transformer conforming to IEC 60601-1 for networks between the computer and HUB. Otherwise, there is a risk of electric shock if the network is faulty.

Do not install in locations exposed to water, steam, moisture or dust.
Doing so may cause problems or malfunctions.
Do not install in locations exposed to salt, sulfur or corrosive gas.
Doing so may result in corrosion of the instrument, problems or malfunctions.
Do not install in locations that are unstable or exposed to vibration.
The vibration may knock over the instrument or the instrument may become unbalanced and fall, resulting in a malfunction or injury.
Do not place anything near the power plug.
To make it easy to unplug the power plug at any time, avoid putting any obstructions near the AC outlet. Fire or electric shock may result if the power plug is not unplugged during an emergency.
Do not block the vent holes.
Doing so may cause the internal temperature to rise and result in fire.
Do not place your hands or fingers on the base.
Your hand or fingers may be pinched and injured when the stage moves to either side. Similarly, instruct the patient not to place his/her hands or fingers on the base.
Do not hold the head rest or digital camera when moving the CR-2.
When carrying the CR-2, be sure to tighten the stage lock, hold the carrying indentations at the front and rear of the base panel, and keep the CR-2 level. Do not hold it by the digital camera or the head rest or any other parts, as they may come off and result in injury.
Hold the CR-2 when connecting or disconnecting a cable.
For safety reasons, when connecting or disconnecting the power cord or any cable, be sure to hold the main unit. Otherwise, the main unit may fall over, possibly causing injury.
Ensure that the entire system conforms to IEC 60601-1-1.
Use the computer, monitor, and other equipment that conform to the system standard IEC 60601-1 or IEC 60950-1 for the Digital Retinal Camera CR-2. Be sure that the entire system conforms to IEC 60601-1-1. Be sure to also use an isolation transformer conforming to IEC 60601-1 when a computer or monitor conforming to IEC 60950-1 is used. Otherwise, electric shock may occur. For details, please contact your sales representative or local Canon dealer.

To prevent the risk of infection, wipe the forehead rest with disinfectant ethanol for each patient. To ensure cleanliness, replace the chin rest paper for each patient. If the chin rest paper is not being used, be sure to disinfect the chin rest for each patient. For details on how to disinfect, consult a specialist. The forehead rest may be corroded if a disinfectant other than those above is used.
Slowly move the main unit towards the patient’s eye when taking an image.
When adjusting the position of the main unit in the front-back direction, slowly bring the main unit closer to the patient while looking at the patient’s eye from the side. The patient’s eye may be injured if the objective lens makes contact with it.
Check the image before using the CR-2.
Before using the CR-2, be sure to take a test image to ensure that there is no foreign matter present that can affect image readings or diagnosis.
Be sure to turn off the power when not in use.
For safety reasons, be sure to turn off the power of all the devices when the CR-2 is not being used. Also, remove the power plug from the AC outlet and put on the cover when the CR-2 is not going to be used for a long time. Otherwise, dust or any foreign matter may accumulate and result in fire.
Do not touch the main unit or the chin rest while they are moving.
The main unit and the chin rest move to the center position when the CR-2 main unit is turned on or a study is finished with the control software. Do not touch the main unit or the chin rest while they are moving. Keep the patient’s chin away from the chin rest.
Before packing the CR-2, return the main unit and chin rest to their initial positions.
Lower the main unit and chin rest to the bottom. To automatically lower them, turn on the power while holding down the set button.
Inspect daily and periodically.
For safety reasons, before using the CR-2, be sure to perform the daily inspection. Have a periodically inspection performed for the CR-2 at least once a year by a Canon designated representative to maintain its performance and reliability.
Do not turn off the CR-2 or the computer during image capture, data transfer, or backup.
Doing so might damage the computer or corrupt the data.
Do not disconnect the cable between the CR-2 and the computer during image capture or data transfer.
Doing so might damage the computer or corrupt the data.

Digital camera
2 Safety

Temperature Humidity Atmospheric pressure
Environment of use 10 to 35°C 30 to 90% RH (no condensation)
700 to 1060 hPa
Storage and transportation environment
–30 to 50°C 10 to 95% RH (no condensation)
700 to 1060 hPa
2 Safety

Disposal of this product in an unlawful manner may have a negative impact on human health and on the environment. Therefore, when disposing of this product, be absolutely certain to follow the procedure which conforms with the laws and regulations applicable to your area.
This symbol indicates that this product is not to be disposed of with your household waste, according to the WEEE Directive (2002/96/EC) and your national law. This product should be handed over to a designated collection point, e.g., on an authorized one-for-one basis when you buy a new similar product or to an authorized collection site for recycling waste electrical and electronic equipment (EEE). Improper handling of this type of waste could have a possible negative impact on the environment and human health due to potentially hazardous substances that are generally associated with EEE. At the same time, your cooperation in the correct disposal of this product will contribute to the effective usage of natural resources. For more information about where you can drop off your waste equipment for recycling, please contact your local city office, waste authority, approved WEEE scheme or your household waste disposal service. For more information regarding return and recycling of WEEE products, please visit www.canon-europe.com/WEEE * EEA: Norway, Iceland and Liechtenstein
2 Safety

The position and contents of the label attached on the CR-2 is shown below. Follow the information on the label to use the CR-2 appropriately.

The following table describes the marks and indications on the rating label.
Class I device indicating manufacturer’s declaration of conformance with the Annex VII of the Medical Device Directive, 93/42/EEC
Certification mark that shows the product has passed the tests by TUV Rheinland of conformity with the medical device regulations in the European Union.
Product that WEEE directive, Directive on Waste Electrical and Electronic Equipment, 2002/96/EC, requires separate collection. The directive is effective in the European Union only.
Certification mark that indicates the product complies with UL 60601- 1 and CAN/CSA 22.2 No. 601.1, that specifies protection against fire, electric shock, and mechanical hazards
Year and month of manufacture Example: October 2011
Serial number in six digits Example: 123456
2 Safety · Alternating current · Type B


1 Digital camera 2 Power switch for the digital camera 3 Digital camera cover 4 Observation monitor
12 Height adjustment mark 13 Main unit 14 Operation panel 15 USB connector 16 AC power connector 17 Power switch for main unit 18 Stage
For viewing the anterior segment and retina.
5 Operation lever 6 Shutter release button 7 Alignment button
Switches between a view of the anterior segment and the retina.
8 Main unit vertical movement ring
Turn right: Main unit moves up.
Turn left: Main unit moves down.
Turn to focus the image.
10 Stage lock 11 Indentation for lifting
9 Focus ring

3 Names of Parts

19 Connector for external eye fixation lamp (with cap)
26 Diopter compensation knob
a No diopter compensation
Connector for the external eye fixation lamp (sold
separately).
20 Forehead rest 21 Objective lens 22 Objective lens cap 23 Head rest 24 Chin rest 25 Indentation for lifting
d Anterior segment photography
b Negative compensation · c Positive compensation · 27 Base

3 Names of Parts

1 Flash intensity button/lamp
6 FIX TARGET button
+: Flash intensity increases.
Move up, down, right, and left to move internal eye fixation lamp.
–: Flash intensity decreases.
2 Observation light brightness button/lamp
+: Observation light brightness increases.
Internal eye fixation lamp blinks.
–: Observation light brightness decreases.
Lights when power is turned on.
3 x2 photography button/lamp
Enables x2 photography function. (See page 30)
4 Small pupil photography button/lamp
Enables small pupil photography function. (See
5 CHIN REST button
Top: Raises chin rest.
Bottom: Lowers chin rest.
7 Set button/lamp · 8 POWER lamp · page 30)

3 Names of Parts

1 Photography ready lamp
4 Left/right eye indicator
R: Right eye
L: Left eye
2 Pupil alignment circles 3 Observation light brightness indicator
5 Diopter compensation knob position indicator
: Anterior segment photography
6 Flash intensity indicator

1 Indicator for top limit of position of main unit
7 x2 photography range 8 Diopter compensation knob position indicator
Appears when main unit has reached top limit.
2 Focus indicator 3 WD guide (green) 4 Indicator for bottom limit of position of main unit
+: Positive compensation
–: Negative compensation
9 Small pupil photography range 10 Internal eye fixation lamp position indicator 11 Focus lens position indicator
Appears when main unit has reached bottom limit.
5 x2 photography indicator 6 Small pupil photography indicator

The flow of operations for retinal photography is as follows. Connect the cables if necessary as described in “Connecting Cables” (see page 22).

Preparing for an Examination (page 22)
Patient Setup (page 24)

Small Pupil Photography Function (page 30)
x2 Photography (page 30)
Diopter Compensation (page 31)
Anterior Segment Photography (page 32)
Finishing an Examination (page 29)
Alignment (page 25) · Photography (page 28)

Connect the CR-2 to a computer in which the supplied software, the Rics NM 2, has been installed.
Otherwise, fire or electric shock may result.
1 Make sure that the power switch for the CR-2 is turned off.
2 Connect the power cord and the USB cable to the retinal camera.
3 Insert the power plug all the way into an AC outlet.
4 Connect the other end of the USB cable to the USB connector of the PC.
Use a high-speed AB type USB cable that is shorter than three meters. For details, contact your sales representative or local Canon dealer.

1 Turn on the computer and start the Rics NM 2.
2 If the digital camera is off, turn the switch to the ON.

If the Mode Dial is set to anything other than , an image cannot be taken. While holding down the button at the center of the Mode Dial, turn the dial to .
4 Basic Operations


3 Remove the objective lens cap and turn the retinal camera's power switch to the ON position.
The POWER lamp lights and the set lamp blinks. The CR-2 automatically does an initialization sequence (internal check and initialization of driving parts). After the initialization the parts of the main unit stop in their center positions.

Power-saving function If the power is on and the buttons on the operation panel are not operated for over 10 minutes, the power-saving function is activated. The POWER lamp blinks while in power saving mode. To return to normal mode, press any button on the operation panel, move the stage right or left, or turn the diopter compensation knob.
4 Make sure the diopter compensation knob is in the 0 position.

You can also check the position of the diopter compensation knob on the observation monitor.
5 Press and hold the unlock button and slide the stage lock to the UNLOCK side.
Align the mark on the stage lock and the mark on the left side of the base. The stage lock is released.
6 Move the stage towards you with the operation lever.
4 Basic Operations

7 Disinfect the forehead rest and replace the chin rest paper.
If chin rest paper is not being used, disinfect the chin rest also.

1 Start an examination on the Rics NM 2.
Enter the patient’s information. The set lamp lights and the observation monitor turns on.
2 Seat and position the patient.

3 Move the chin rest with the CHIN REST button so that the patient's eye is aligned with the height adjustment mark.
4 Move the stage toward the eye to be photographed.

Otherwise, your or the patient’s hands or fingers may be pinched, possibly resulting in injury.
Hold the operation lever, pull it toward you, and then move the stage.
4 Basic Operations · Chin rest rises · Chin rest lowers

1 Align the retinal camera with the eye to be photographed.
During examination of the anterior segment of a patient who is extremely near- or far-sighted, the patient may not be able to see the internal eye fixation lamp.
When the vertical position cannot be fully adjusted by using the operation lever Adjust the height of the chin rest with the CHIN REST button to raise or lower the patient's face.
4 Basic Operations · Pupil alignment circles

2 Tilt the operation lever forward or backward to unify the top and bottom of the image of the pupil.
When the pupil is smaller than the inner pupil alignment circle, images cannot be taken properly. Darken the room further to dilate the patient's pupil more. If the patient's pupil does not dilate enough, use the small pupil photography function (see page 30).
3 Press the alignment button.
The retina appears in the observation monitor and the internal eye fixation lamp blinks for three seconds.
4 Adjust the observation light brightness with the observation light brightness button.
Guidelines for observation light brightness Adjust to the lowest possible brightness level to reduce the strain on the patient's eye.
Observation monitor brightness adjustment Use the Rics NM 2 to adjust the brightness of the observation monitor. See the Rics NM 2 manual for details.
4 Basic Operations

5 Have the patient stare at the internal eye fixation lamp to fix the part for the photograph.
Take the following steps to guide the patient’s eye: Press the FIX TARGET buttons up, down, left, and right to move the internal eye fixation lamp to your desired position and then press the set button to make it blink. If the eye cannot be guided by the internal eye fixation lamp, use the external eye fixation lamp (sold separately).
Press the set button for at least two seconds to reset the position of the internal eye fixation lamp. See the Rics NM 2 manual for details.
6 Turn the focus ring to align the split lines into a single line.
The image enlarges while the focus ring is being turned.
When the split lines do not align If the patient's diopter range is outside the –10 to +15 D, it is not possible to focus with the focus indicator. If this happens, use the diopter compensation function (see page 31), and focus the image until it is sharp.

1 Do a preliminary check before taking photos.
a. The photography ready lamp lights green. b. The working distance dots appear sharpest. c. The working distance dots are on the extension line of the WD guides. d. The split lines form a single line.

Photography ready lamp • Green: Ready to shoot.
Flash intensity adjustment The flash intensity is set in advance to the standard level. Every time the flash intensity button is pressed the flash intensity increases by 0.3 stop.
Appearance of flare Even if the working distance dots are adjusted to be extremely clear, flare may appear around the retinal image depending on the patient’s eye.
2 Press the shutter release button.
The image of the retina and the examination information appear on the Rics NM 2 screen.
3 Check the photographed image.
Check that the image is normal and there are no errors in the examination information. To take images of the other eye, repeat beginning with step 1 of Alignment (see page 25).
If flare appears See “7 Troubleshooting” (see page 40) and follow the procedure.

1 Save the images on the Rics NM 2 and end the examination.
The set lamp starts blinking and the observation monitor turns off.
2 Align the stage with the base and slide the stage lock to LOCK side to lock the stage.
Align the mark on the stage lock and the mark on the right side of the base.
3 Turn off the power to the CR-2.
4 Exit the Rics NM 2 and turn off the computer.
4 Basic Operations

The section describes how to use x2 photography, small pupil photography function, diopter compensation, and anterior segment photography. Instructions are also included for connecting an optional external monitor.

You can take and save images that are within the x2 photography range on the observation monitor.
The lamp on the x2 photography button lights, and the x2 photography indicator and x2 photography range marks appear on the observation monitor.
The image in the x2 photography range appears on the Rics NM 2 screen.
If the patient’s pupil does not dilate enough, use the small pupil photography function. When the function is used, pupil diameters 3.3 mm or larger can be photographed.
1 Press the small pupil photography button.

The lamp on the small pupil photography button lights, and the small pupil photography indicator and a dotted circle appear on the observation monitor.
Small pupil photography indicator

5 Photography Auxiliary Functions
2 Take the image.
Press the small pupil photography button again to finish the small pupil photography.
If because of the patient's diopter the image cannot be brought into focus by turning the focus ring, use the diopter compensation knob to change the position. Insert the diopter compensation lens into the CR-2 and take the photograph.
The CR-2 can provide appropriate images if the diopter of the patient’s eye is within the respective ranges below.
When operating the diopter compensation knob, set it precisely to the 0, –, or + position. If the knob is not aligned with the position noted above, the retinal image is not captured correctly.
Diopter compensation mark · Negative · Positive

5 Photography Auxiliary Functions
This function makes it possible to photograph the anterior segment. It is also possible to use the x2 photography function at the same time.
Do not allow external light that contains infrared to enter the position being photographed.
1 Switch the diopter compensation knob to the anterior segment photography position.
The anterior segment photography mark appears on the observation monitor.
Anterior segment photography mark
2 Move the operation lever to align the area to be photographed in the center of the observation monitor.
3 Tilt the operation lever to the front or back to focus on the area to be photographed.
4 Press the shutter release button.
The image of the retina appears on the Rics NM 2 screen. To take images of another position, repeat steps from step 2.
The image on the observation monitor can be viewed on an external monitor. Contact your sales representative or local Canon dealer about which external monitors and HDMI mini cable can be used.
Anterior segment photography

5 Photography Auxiliary Functions
1 Remove the digital camera cover.
2 Connect the HDMI cable to the digital camera.

4 Connect the HDMI cable to the external monitor.
Connect the HDMI cable to the HDMI input terminal on the external monitor.
5 Turn on the power to the retinal camera and the digital camera.
The image is displayed on the external observation monitor. No image is displayed on the observation monitor of the retinal camera.
Notch

This section outlines the maintenance instructions for the CR-2: daily inspection, cleaning, disinfecting, refilling chin rest paper, and removing/installing the digital camera.
Perform the following inspections before using the CR-2 to insure that it is used safely and correctly. If a problem is found during the inspection and you are unable to correct the problem, please contact your sales representative or local Canon dealer.
Check the following items before turning on the power.
Turn on the CR-2 and start the Rics NM 2. Check the following items after starting an examination.

For information about lens cleaning paper, lens cleaner, and blowers, contact your sales representative or local Canon dealer.
If the objective lens is dirty, clean it according to the procedure below.
1 Check for any dirt.
Illuminate the objective lens with a penlight to check for dirt.
2 Blow away any dust or dirt.
Use a blower to blow away any dust or dirt on the lens. Do not use a brush to dust off.
Do not allow the blower to touch the objective lens.
3 Wipe the objective lens.
Use lens cleaning paper and lens cleaner designated by Canon.
6 Maintenance

The digital camera has a self-cleaning sensor unit. Dirt and dust on the front surface of the imaging sensor are automatically removed when the power is turned on or off. This function makes it possible to remove the majority of dirt and dust that may interfere with images. It is not necessary to be aware of this function during normal usage.
To use the self-cleaning sensor unit for cleaning, follow the procedure below.
1 Make sure that the power switch for the retinal camera is turned off.
2 Switch the digital camera's power switch from on to off.
The self-cleaning sensor unit starts and the imaging sensor is cleaned.
3 Check that the dirt and dust have been removed.
4 Switch the digital camera's power switch to on, photograph an image and check the image.
If the above procedure does not remove the foreign object, the cause may not be dirt or dust. Contact your sales representative or local Canon dealer.
Otherwise, fire or electric shock may result.
Do not clean the outside of the CR-2 with lens cleaner. It may corrode the material of the outside of the CR-2.
If the outside of the CR-2 is dirty, wipe it twice with a soft cloth. First use a cloth that has been soaked in diluted neutral cleanser and well wrung out. Then use a cloth that has been soaked in water and well wrung out.
6 Maintenance

For details on how to disinfect, consult a specialist.
Use a sanitized gauze or wipe that includes disinfectant to clean the forehead rest for each patient.
1 Pull out the right and left chin rest holding pins.
2 Insert the holding pins into the holes on the right and left of the chin rest paper.
3 Attach the chin rest paper to the chin rest with the pins pressing into the holes on the chin rest.
Chin rest paper is a consumable product (sold separately). To purchase chin rest paper, contact your sales representative or local Canon dealer.
The CR-2 is equipped with a specialized digital camera. It is not necessary to attach or remove the digital camera during normal usage. To replace the digital camera, follow the procedure below to remove and attach the digital camera.
Although it is not recommended, if you need to remove and attach the digital camera, take the following precautions.
6 Maintenance

1 Turn off the power to the retinal camera and remove the plug from the AC outlet.
2 Loosen the cover screw and remove the digital camera cover.

3 Remove the cables from the digital terminal and the remote control terminal.
4 Remove the DC coupler.
5 While holding down the lens release button on the digital camera, turn the digital camera counterclockwise to remove it.
6 Attach the caps to the retinal camera and the digital camera mount.
6 Maintenance

1 First be sure to check the following items.
2 Attach the digital camera to the retinal camera.

3 Attach the DC coupler to the digital camera.
4 Open the terminal cover of the digital camera and attach the cables to the digital terminal and the remote control terminal.
5 Attach the digital camera cover to the digital camera.
6 Maintenance

If any of the symptoms listed here occurs or if a warning appears in the Rics NM 2 while using the CR-2, try the relevant remedy described below. If these remedies do not eliminate the symptoms or the warning, turn off the power and contact your sales representative or local Canon dealer. Be ready to provide a detailed description of the symptoms and message.
Symptom Cause and remedy
The power cord has been disconnected. Turn off the power to the retinal camera and reconnect the power cord.
The power lamp does not light when the power is turned on.
The power cord has been cut. Contact your sales representative or local Canon dealer.
5 short beeps sound and the operation panel lamps blink when the power is turned on.
A system error has occurred. Contact your sales representative or local Canon dealer.
The observation monitor is dim and hard to see.
Adjust the brightness of the observation monitor. See the Rics NM 2 manual.
The observation monitor does not turn on. Turn on the digital camera.
An error appears on the observation monitor.
A malfunction has been detected in the digital camera. Turn off the power to the retinal camera and then turn it on again.
Err 06 Sensor cleaning is not possible.
Err 70 Photography is not possible due to an error.
The working distance dots are hard to see. The observation light is too bright. Reduce the observation light brightness. The split lines are hard to see.
The working distance dots are not visible.
The retinal camera and the eye are not aligned. Align them correctly.
The retinal camera and the eye are not aligned. Align them correctly. If flare still appears, refer to items 1 through 4 below and try taking the photo again.
Flare appears in the image.
The retinal camera and the patient’s eye are separated. Move the objective lens closer to the patient’s eye.
The retinal camera and the patient’s eye are too close. Move the objective lens away from the patient’s eye.
The retinal camera is too close to the left side of the patient’s eye. Move the objective lens to the right.
The retinal camera is too close to the right side of the patient’s eye. Move the objective lens to the left.
The patient's pupil is not dilated enough. Dilate the patient’s pupil. If the pupil is still too small, enable the small pupil photography function.
Eyebrows or eyelashes are blocking the pupil. Instruct the patient to open his/her eye wider.
The split lines are not visible.
The patient's eye is cloudy. Focusing cannot be performed with the split lines.
The diopter compensation knob is set to either negative or positive. Set it to the 0 position.
The split line lamps are not lit.

Symptom Cause and remedy
The diopter patient's eye is outside the range of –10 to +15 D. Use the diopter compensation function, turn the focus ring, and focus on the position where the images of the retina look clear at the same time.
The split lines do not align.
The patient's pupil is not dilated enough. Dilate the patient’s pupil. If the pupil is still too small, enable the small pupil photography function.
The retinal camera and the eye are not aligned. Align them correctly.
Focus the image.
Eyebrows or eyelashes are blocking the pupil. Instruct the patient to open his/her eye wider.
Cannot capture good images.
The patient's eye is cloudy. Clear images cannot be captured if this is the case.
Adjust the flash intensity, patient's diopter, and small pupil photography function.
The objective lens is dirty. Clean the objective lens (see page 35).
The patient is wearing contact lenses. Instruct the patient to remove his/her contact lenses.
There is dirt or dust on the objective lens. Clean the objective lens (see page 35).
Black spots or any artifacts appear in the same place in every captured image.
Dirt or dust is stuck to the imaging sensor of the digital camera. Use the self-cleaning sensor unit to clean it. (see page 36).
The objective lens is dirty. Clean the objective lens (see page 35).
White spots are visible on the captured image.
The patient's eyelashes are interfering with the image. Instruct the patient to open his/her eye wider, and capture another image.
7 Troubleshooting

Photographic magnification 1.1 x (image size on the sensor)
Required pupil diameter 4.0 mm or greater (3.3 mm when using small pupil photography function)
Working distance 35 mm
Focus adjustment range –10 to +15 D (without compensation lens) –31 to –7 D (when using negative compensation lens) +11 to +33 D (when using positive compensation lens) (Split-line alignment type focusing) * 1 D = 1 m–1
Flash intensity Linked to the photography mode Can be set manually
Observation light source Infrared LED
Flash source White LED
Eye fixation lamp Internal eye fixation lamp (LED) External eye fixation lamp (sold separately)
Stage 100 mm side to side, 70 mm front to back
Main unit vertical movement
Chin rest range of motion 60 mm
Power supply rating AC 100 V to 240 V, 50/60 Hz, 1 A to 0.4 A
Dimensions 305 (W) × 500 (D) × 473 (H) mm
Weight Approximately 15 kg (including 0.8 kg digital camera)

The CR-2 is designed and tested to comply with IEC 60601-1-2 (EN 60601-1-2), the applicable regulations regarding EMC for medical devices. The CR-2 needs to be installed and put into service according to the EMC information stated as follows.
Name Type Length Remarks
AC Power Cord Non-Shielded 3.0 m fixed-length Supplied
USB Cable Shielded Max. 3.0 m Not supplied
HDMI Cable Shielded Max. 2.7 m Not supplied
4. The use of accessories, transducers and cables other than those specified, with the exception of transducers and cables sold by Canon sales representative or local Canon dealer as replacement parts for internal components, may result in increased emissions or decreased immunity of the CR-2. 5. The CR-2 should not be used adjacent to or stacked with other equipment and if adjacent or stacked use is necessary, the CR-2 should be observed to verify normal operation in the configuration in which it will be used.
Appendix

The CR-2 is intended for use in the electromagnetic environment specified below. The customer or the user of the CR-2 should assure that it is used in such an environment.
Emission Test Compliance Electromagnetic Environment – Guidance
The CR-2 uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electromagnetic equipment.
CISPR11
EN 55011
CISPR11
The CR-2 is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
EN IEC 61000-3-2 Class A
emissions EN IEC 61000-3-3*
* Not applicable to regions where the rated voltage is less than 220 V.

The CR-2 is intended for use in the electromagnetic environment specified below. The customer or the user of the CR-2 should assure that it is used in such an environment.
Immunity Test IEC 60601
Test Level Compliance Level Electromagnetic Environment – Guidance
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Electrostatic discharge (ESD) IEC 61000-4-2
±(2, 4, 6) kV contact
±(2, 4, 6) kV contact
±(2, 4, 8) kV air
±(2, 4, 8) kV air
±2 kV for power supply lines
±2 kV for power supply lines
Electrical fast transient/ burst IEC 61000-4-4
Mains power quality should be that of a typical commercial or hospital environment.
±1 kV for input/output lines
±1 kV for input/output lines
±1 kV differential mode
±1 kV differential mode
Mains power quality should be that of a typical commercial or hospital environment.
±2 kV common mode
±2 kV common mode
95% dip in UT) for 0.5 cycles
95% dip in UT) for 0.5 cycles
Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11
Mains power quality should be that of a typical commercial or hospital environment. If the user of the CR-2 requires continued operation during power mains interruptions, it is recommended that the CR-2 be powered from an uninterruptible power supply.
40% UT (60% dip in UT) for 5 cycles
40% UT (60% dip in UT) for 5 cycles
70% UT (30% dip in UT) for 25 cycles
70% UT (30% dip in UT) for 25 cycles
95% dip in UT) for 5 sec
95% dip in UT) for 5 sec
Power frequency (50/60Hz) magnetic field IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
Appendix · Surge IEC 61000-4-5

The CR-2 is intended for use in the electromagnetic environment specified below. The customer or the user of the CR-2 should assure that it is used in such an environment.
Immunity Test IEC 60601
Test Level Compliance Level Electromagnetic Environment – Guidance
Portable and mobile RF communications equipment should be used no closer to any part of the CR-2, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter
Recommended Separation Distances d = 1.2 √P d = 1.2 √P 80 MHz to 800 MHz d = 2.3 √P 800 MHz to 2.5 GHz
Conducted RF IEC 61000-4-6
3 Vrms 150 kHz to 80 MHz
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m).
Radiated RF IEC 61000-4-3
3 V/m 80 MHz to 2.5 GHz
Field strengths from fixed RF transmitters, as determined by an electromagnetic site surveya, should be less than the compliance level in each frequency rangeb.
Interference may occur in the vicinity of equipment marked with the following symbol:
reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the CR-2 is used exceeds the applicable RF compliance level above, the CR-2 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the CR-2. b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Appendix · 3 Vrms · 3 V/m

Recommended Separation Distances between Portable and Mobile RF Communications Equipment and Devices or the CR-2
The CR-2 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the CR-2 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the CR-2 as recommended below, according to the maximum output power of the communications equipment.
Separation distance according to frequency of transmitter
150 kHz to 80 MHz d = 1.2 √P
80 MHz to 800 MHz d = 1.2 √P
800 MHz to 2.5 GHz d = 2.3 √P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.

Service life
The service life of this product is eight years if specified inspections and maintenance are done.
If a problem cannot be solved even after taking the measures indicated in “7 Troubleshooting” (see page 40), contact your sales representative or local Canon dealer for repairs. When requesting repair, please provide us with the following information.
Name of the instrument: CR-2 Serial number: 6-digit number on the rating label Description of malfunction: Report as much detail as possible.
Performance parts (parts for repairs to maintain performance) will be stocked for eight years after production of the CR-2 is discontinued.
The following parts are consumable or may deteriorate because of characteristics of their material or construction. These parts cannot be replaced by the user. If these parts are found to be worn out or to have deteriorated during daily or regular inspection, contact your sales representative or local Canon dealer for repairs.
Appendix







CANON INC. Medical Equipment Group 30-2, Shimomaruko 3-chome, Ohta-ku, Tokyo 146-8501, Japan Telephone: (81)-3-3758-2111
Distributed by: CANON U.S.A., INC. Healthcare Solutions Division 15955 Alton Parkway, Irvine, CA 92618, U.S.A. Telephone: (1)-949-753-4160
CANON EUROPA N.V. Medical Products Division Bovenkerkerweg 59, 1185 XB Amstelveen, The Netherlands Telephone: (31)-20-545-8926
CANON SINGAPORE PTE. LTD. Medical Equipment Products Division 1 HarbourFront Avenue, #04-01 Keppel Bay Tower, Singapore 098632 Telephone: (65)-6799-8888
CANON AUSTRALIA PTY. LTD. Optical Products Division 1 Thomas Holt Drive, North Ryde, Sydney N.S.W. 2113, Australia Telephone: (61)-2-9805-2000
L-IE-5176C 0613P0.001 © CANON INC. 2012 Printed in Japan
Owners and our team usually reply within a day — and your question helps the next person too.
Be the first to ask — other owners or our team will answer.