AU2700 & AU5400 ON-LINE SPECIFICATIONS

Online Specifications
Olympus Optical Co., Ltd.


(1) Test Requisition Information (T.R.I.) (2) Test Results
(1) Class A Transmission is without synchronizing with analysis. This mode does not utilize the ACK/NAK protocol. (2) Class B Transmission is synchronizing with analysis. This mode does utilize the ACK/NAK protocol.
(1) Setup (2) Communication protocol
A. T.R.I. Inquiry Text Format B. Test Result Text Format (3) Test Number Assignment

(2.1) Format of Transmission



(2.3) Text format (1) Basic configuration as follows:
remarks * _ is space.

A. Definition of Term
(2) Blocking

B. Kind of text and blocking
Remark *1:When Block Identification No. is numeric(0,1, ), Only “End of Block” is used.
When Block Identification No. is “E”, Only “End of Text” is used.

(3.1) Common and changeable item
The following can be selected by [PARAMETER] [ONLINE] [PROTOCOL] menu. Text start code 1) Unit No. 3) Text end code 5) BCC 6)
Sample Type Dummy(in Test Result transmission Text) Block Identification No. Rack No./Cup Position Unit No. on the T.R.I. text (with out Rerun) ANL unit No. and Cuvette side on Test Result transmission Text (without RB,ACAL,QC) Space Code before the Text Classification (S,R,D text) Reagent Lot No. and bottle No.
The order of the Sample number and Sample ID on the text can be changed, the sample type, the Dummy spaces and the Block Identification Number can be changed in the [PARAMETER] [ONLINE] [PROTOCOL] menu. There is a function at the bottom of the page called “Special”. These features may be altered under the “Special” menu. Making changes in the “Special” menu only affects the T.R.I. Text and Test Result Transmission Text.
(3.2) T.R.I. inquiry text
(1) Start text to inquiry test requisition.
(2) Normal run inquiry test requisition.

Common Item

(3) Repeat run inquiry test requisition.

(4) End text to inquiry test requisition.
Sample type

(3.3) T.R.I. text
(1) Normal sample test requisition information text.

Sample type Block Identification No.
Test code 5) 6)
(2) Repeat run sample test requisition information text.

Original Sample no.
Sample type Block Identification No.


(3) End of T.R.I. Text transmission.
(3.4) Test result transmission text
(1) Start of test result transmission text.

(2) Normal sample test result text.

Sample no. Sample ID Dummy
Sample type Block Identification No.


(3) Stat(fast) test result text (AU2700 only).

P n n n Sample no. Sample ID Dummy
1-22 Sample type Block Identification No.


(4) Emergency(fast) test result text (AU2700 only).

E n n n Sample no. Sample ID Dummy
1-22 Sample type Block Identification No.


(5) Repeat run result text -1 .

Original Sample no.
Sample type Block Identification No.

Blocking · Blocking

(6) Repeat run result text -2 Stat fast sample (AU2700 Only) .

Original Sample No.
1-22 Sample type Block Identification No.

Blocking · Blocking

(7) Reagent blank result text.

This dummy is same ID digits Block Identification No. and assigned area is decided associated with the rack No. ANL unit No. Cuvette side
Test code R1 Lot No. R1 Bottle No. R2 Lot No. R2 Bottle No.

(8) Acal (Calibration) result text.

This dummy spaces is assigned in case rack No. exists Calibrator No. (1-80) and assigned area is decided associated with the rack No. Block Identification No.

Blocking · Blocking

(9) QC (Quality Control) result text.

This dummy No. is assigned in case rack No. exists Control No. and assigned area is decided associated with the rack No. Block Identification No. ANL unit No.

(10) End of test result transfer text.
Blocking · Blocking

(3.5) Relation between real-time/batch and Transmission text

(3.6) Others
(1) Contents and formats of text.
Item Digit Contents Remarks 1. Rack No.
4 or 5 '0001'-'9999' or '00001'-'99999' STAT sample is space
When Rack no. is no setting, Rack No. is '00000'.
For Repeat Run Results.
(Optional)

9.ANL unit No.
ANL unit No. is added to the header portion of RB, ACAL, and QC result data. Normal sample and rerun sample, it is added to each item. However, in the case of an calculate item, a space is added. In a normal sample and repeat run, it can be prevented from adding a unit number by the setup of an on-line condition menu. This option can be selected in the [PARAMETER][ONLINE] [PROTOCOL][SPECIAL] menu. 10.Cuvette side
Cuvette side is added to the header portion of RB, ACAL, and QC result data. Normal sample and rerun sample, it is added to each item. However, in the case of an calculate item, a space is added. In a normal sample and rerun, it can be prevented from adding a unit number by the setup of an on-line condition menu. This option can be selected in the [PARAMETER][ONLINE] [PROTOCOL][SPECIAL] menu. 11. Sex
1 AU5400 ‘0’ : inner cuvette (ISE Cell 1) ‘1’ : outer cuvette (ISE Cell 2) AU2700 ‘0’ : inner cuvette(ISE) ‘1’ : outer cuvette
20 number or character
2 '01'-'99' • This number can be changed in the [PARAMETER] [ONLINE] [TEST NO.] 16. Method 1 '0':normal '1':dilution ‘2’:condense
For Repeat Run Results.
(Optional) · (Optional) · (Optional)

Refer to Appendix 1 (Flags List)
4*4 Number or character • This option can be selected in the [PARAMETER] [ONLINE] [PROTOCOL] [SPECIAL] menu.
Remark *1: If only age is the space in the text from a host, AU2700/AU5400 will output an error. The combination of Age
and Month is as follows.
Pattern Age Month Result 1 Spaces Spaces OK 2 XXX Spaces OK 3 Spaces XX Error 4 XXX XX OK
Host -> AU2700/5400

(2) Test result format.
A. Test result
0 1 2 3 . 4
- 0 1 2 3 . 4 5 6
- _ 1 2 3 4
- _ 1 2 3 . 4 5 6

B. LIH result
0 NORMAL FLAG 1 + 2 ++ 3 +++ 4 ++++ 5 +++++ 6 ABN 9 Test not performed.
0
0 NORMAL FLAG 1 + 2 ++ 3 +++ 4 ++++ 5 +++++ 6 ABN 9 Test not performed.

0 NORMAL FLAG 1 + 2 ++ 3 +++ 4 ++++ 5 +++++ 6 ABN 9 Test not performed.
0 NORMAL FLAG 1 + 2 ++ 3 +++ 4 ++++ 5 +++++ 6 ABN 9 Test not performed.
_ _ _

(3) T.R.I. of calculated tests.
(4) Data format that Test result data is digit over.
a. Data digit is 6.
9
9
(5) Data format that Test result data is OD-value.
When the data is negative OD-value, it is same as digit over.
- 0 . 1 2 3 4

(6) T.R.I of LIH.
In any case, LIH test no. in T.R.I from Host is necessary for LIH analysis.
LIH Reagent Test Name LIH Measure All Even if AU2700/AU5400 does not have the request from a host, it performs LIH analysis. LIH
Select AU2700/AU5400 performs LIH analysis according to the request from a host. All Even if AU2700/AU5400 does not have the request from a host, it performs LIH analysis. Other
No LIH analysis is not performed even if AU2700/AU5400 has the request from a host. Note 1) These can be selected by the following menus.
LIH Reagent Test Name : [PARAMETER] – [COMMON TEST PARAMETERS] – [TEST NAME] LIH Measure : [PARAMETER] – [COMMON TEST PARAMETERS] – [ROUND]

(4.1) Protocols at AU2700/5400
(1) Class A: without ACK/NAK exchange.
A. Sending
Text to be sent:
T.R.I. Inquiry text Repeat Run T.R.I. Inquiry text
Start of Test Result Transmission text Test Result text Repeat Run Result text End of Test Result Transmission text
Start Block end (BCC) Initial of Transmission text Condition Text Text end (BCC)
B. Receiving
Text to be received:
T.R.I. text
Repeat Run T.R.I. text End of T.R.I. Transmission text
Start Block end (BCC) Initial of Transmission text Condition Text Text end (BCC)

(2) Class B: with ACK/NAK protocol.
A. Sending
Text to be sent:
T.R.I. Inquiry text Repeat Run T.R.I. Inquiry text Start of Test Result Transmission text Test Result text Repeat Run Result text End of Test Result Transmission text
ACK ACK Received Received
Block end (BCC) Initial Start Condition of Transmission text Text Text end (BCC)
NAK NAK Received Received 1st time: resending the same block or or 2nd time: sending the next block Time-out Time-out

B. Receiving
Test to be sent:
T.R.I. text Repeat Run T.R.I. text End of T.R.I. Transmission text
ACK ACK Send Send
Block end (BCC) Initial Start Condition of Transmission text Text Text end (BCC)
NAK NAK Send Send 1st time: receiving the same block 2nd time: receiving the next block

(4.2) Timing/time-out
(1) Time-out/timing list .
Kind Description Default Remarks T1 Limit time from end of transfer/receive to start of text receive. 2 s 0.1 x n sec T2 Limit time from start of text receive to end of text receive. (Note 1)
T3 Limit time from start of text transfer to end of text transfer. n=1-99 T4 Limit time from end of text transfer to response of receive. 2 s T5 Interval time of transmission 2 s T6 Min. time from receive 1 s T7 Limit time from NAK response to start receive of retry text.
t1 Min. time from end of transfer text to receivable 0.5 s Not Changeable text. t2 Min. time from end of transfer text to next receivable text. 0.5 s
Min. time from end of transfer text to receivable Response. 0.5 s
Min. time from end of receiving text to transfer Response. 0.5 s
(2) Rule of class A

B. Case 2
AU2700/5400
<T3 <T3 <T3
T5< T5<

C. Case 3
AU2700/5400 Host A B C
t2

(4.3) Rule of class B

t3


(4) Case 4 (Abnormal Case 1)

t3

(5) Case 5 (Abnormal Case 2)

t3
(6) Case 6 (Abnormal Case 3)

t4< t2
(7) Case 7 (Abnormal Case 4)

t4< t2

(1) Trans./receive sequence in one session.
General sequence :
A. Example 1 Shift next Rx-transfer when Online Error occurred.

B. Example 2 Stop the session when Online Error occurred.

(2) Detail T.R.I. Receive Kind of text Timing/condition Real-time RB(request This text is transferred when it is shifted from stand-by Start) mode to start measure and then the first Rx is transmitted.
R_ * (request normal) (Note 2) Request of When detecting cup, this Sample No. is transferred if this sample no. sample has not T.R.I.. Request of When reading ID is normally, this Sample ID is transferred sample id if this sample has not T.R.I.. RH (request rerun) (Note 2) Request of When detecting cup, this Sample No. is always transmitted sample no. Request of When reading ID is normally, this Sample ID is always sample ID transmitted. remarks * _ is space. Note 1) The number of times a T.R.I. inquiry can be sent when an online error is generated can be selected in [PARAMETER] [ONLINE] [PROTOCOL]. Note 2) It is possible that the R_-text and RH-text are mixed in one session.

Receive Kind of text Timing/condition
This text can be received after end of Rx-transfer within the designated time. SE RE (request This text is transferred when shifting to the modes below. end) 1) measure mode to standby mode 2) measure mode to stop mode
This text is transmitted when the communication is canceled by the online alarm. Batch RB (request This text is transmitted when receiving T.R.I. is sent in the test start) requisition.
R_ * (request Normal) (Note 3) Request of This text is transmitted at the designated time intervals. sample no. The texts are the samples that are selected in the test requisition menu.
RH (request rerun) (Note 3)
Request of This text is transmitted at the designated time intervals. sample no. The texts are the samples that are selected in the test requisition menu. Sx This text can be received after end of Rx-transfer within the designated time. SE
RE (request end) This text is transmitted when it was receiving the last sample No. in test requisition menu and SE text isn't received. This text is transmitted when the communication is canceled in the test requisition menu. This text is transmitted when the communication is canceled by the online alarm. remarks * _ is space. Note 3) R_-text and RH-text is transferred in separate sessions.
Real-time Sx

(3) Kind of T.R.I. transfer/receive.
A. T.R.I. of normal sample

Setting of parameters Method of T.R.I. inquiry & T.R.I. Receive T.R.I.. inquiry Method of T.R.I. inquiry
Sample no (Calculated rack No. and cup pos. No.) Sample ID Request of sample ID
Sample ID *1 Sample no *1 Batch Sequential Rack no.
(4) T.R.I. of Repeat sample

Setting of parameters Method of T.R.I. inquiry & T.R.I. Repeat sample Receive of T.R.I. Repeat rack
Original sample no. *1 Repeat sample no.*1 Sample ID Request of sample ID
Remark *1: Repeat sample no. and sample ID of T.R.I. must be same as the Repeat sample no. and sample ID of T.R.I
inquiry.

(5.2) Sending test result protocol
(1) Transfer sequence for one session.
General sequence :
A. Example 1
Shift next Dx-transfer when Online Error occurred.

B. Example 2
Stop session when Online Error occurred.

(2) Detail Data Trans. Kind of text Timing/condition Normal end Online error Real DB (START) This text is transmitted when It is shifting the It is output the -time it is shifted from stand-by transfer of the alarm and it is mode to measure mode and just test results. executed the below behind the first Dx is transmitted.
DE (end) This text is transmitted after It is ended the It is output the the last transfer of Dx text session of transfer of alarm and ended and the below shifting mode. The text of test the session of the results. transfer of the 1) measure mode to standby text of test results. 2) measure mode to stop This text is transmitted when the communication is canceled by the online alarm.
2 blocks Host · Host

Data Trans. Kind of text Timing/condition Normal end Online error Batch DB (start) This text is transmitted when it It is shifting the It is output the selected the start of transmission transfer of the test alarm and it is in the online menu. Results. executed the below
DE (end) This text is transmitted when It is ended the It is output the it was transmitting the session of transfer alarm and ended sample No. that is selected in of the text of test The session of the online menu and it was after results. transfer of the the designated time. text of test results.
This text is transmitted when the communication is canceled in the online menu. This text is transmitted when the communication is canceled by the online alarm.
[PARAMETER] [ONLINE] [PROTOCOL] menu.

(1) Mixing the session of the TRT inquiry and the session of the test results.
A. General sequence
Rx Dx Dx Rx
Sx
Section Definition of section Limitation (a) The start of the transmission of the text of T.R.I. inquiry -- The end of the receiving of all T.R.I. for one samples
Don't transmit the Test result text.
(b) The start of the transmission of the text of the test results -- The end of the receiving of all result for one samples
Don't transmit the T.R.I. text.

(6.1) In/Output signals and Terminals
PC Terminal Terminal at AU2700/5400 at host Signal Name Abbrv. Direction computer
(25 Pin Configuration) Ground FG - 1 Data sent TxDATA 3 3 (Rx DATA) Data receive RxDATA 2 2 (Tx DATA) Signal Ground SG 5 7 (SG)
Request to send RTS 7 Clear to send CTS 8
(6.2) Signal Level
Signal/Signal Format Signal Level
SPACE (ON) +3V and higher ("H") MARK (OFF) -3V and lower ("L")
(6.3) Applicable Connector
Style of connector : Confirm the standard of IBM PC (RS232 DB9 Male) Provide a RS232 DB9 Female connector. Style of connector Cover : ditto
(6.4) Others
A. When powering ON or OFF the host computer system, ensure that the AU2700/5400 is in Standy mode. Powering ON or OFF the host computer system while the AU2700/5400 is being operated may result in abnormality function of the AU2700/5400.
B. The cable for host communication is not included in the AU2700/5400 standard accessories. A shielded cable must be used for host connection.
C. Do not use the terminals on the AU2700/5400 other than No. 2,3,5,7, or 8 for any purpose.

A. Appendix
(1) Flags list Flag Contents ( Failed to evade the contamination since the detergent for evading the contamination is not enough. % The sample probe is clogged with sample. ? Calculation unable due to abnormal photometric data
# Sample level detection error R Reagent level detection error ! Calculation unable due to abnormal data
Reagent absorbance value at P-START of Reagent Blank run, is smaller than the Lower limit of the Parameter.
Reagent absorbance value at P-START is lower than the lower limit specified In the Parameters in routine run.
Reagent absorbance value at P-END of Reagent Blank run, is greater than the Upper limit of the Parameter.
Reagent absorbance value at P-END is higher than the upper limit specified In the Parameters in routine run. @ Abnormally high result; absorbance of every wavelength is more than 2.5. $ No linearity validation conducted because less than 3 data obtained in the Kinetics. D Too quick reaction slope in increasing kinetics, absorbance at P-START is Higher than MAX. OD in increasing FIXED assay, or too slow reaction slope In decreasing kinetics (= no reaction observed) B Too quick reaction slope in increasing kinetics, or absorbance at P-END is Lower than MIN. OD in increasing FIXED assay. Linearity error in kinetics. * Linearity error in kinetics. & Prozone data error (prozone check point data error) Z Prozone error F Result higher than the dynamic range specified in the Parameters G Result lower than the dynamic range specified in the Parameters x Other QC data error 1 QC result beyond the QC value range specified in the Parameters 2 Multi-rule QC data error 3 Multi-rule QC data error 4 Multi-rule QC data error 5 Multi-rule QC data error 6 Multi-rule QC data error 7 Multi-rule QC data error p Result beyond the panic value specified in the Parameters T Abnormality found in the Inter-Item Check P Result higher than DECIDE RANGE designated in parameters. N Result lower than DECIDE RANGE designated in parameters. H Higher than the result value range specified in the Parameters L Lower than the result value range specified in the Parameters J Result higher than the repeat run range specified in the Parameters K Result lower than the repeat run range specified in the Parameters S Result extracted for repeat run
u · y

/ Test not performed: test has been requisitioned but not performed due To any reason. r transferred data to host e edited data c correct data by manual correction a On-board stability over b Calibration stability over or Calibration does not exist. ) Current reagent lot number was different from reagent lot No. when ACAL analysis

(2) Online parameters list Transmission information Contents menu T.R.I. receive

Normal Repeat sample Emergency Repeat sample Stat Repeat sample
Normal sample Emergency sample Stat sample Stat Fast sample
Real-time/batch/none Real-time/batch/none Real-time/batch/none Real-time/none Normal Repeat sample Emergency Repeat sample Stat Repeat sample
Real-time/batch/none Real-time/batch/none Real-time/batch/none QC sample Calibration sample Reagent blank sample
Real-time/batch/none Real-time/batch/none Real-time/batch/none Protocol 2 Contents menu Error control
[PARAMETER] [ONLINE] [PROTOCOL] menu. Protocol –1 Contents menu Data format
start code 1 start code 2 end code 1 end code 2 Max text length unit no. use ETB
4800/9600 class A/class B use/none (0-3) nn Text Format
bit/sec protocol BCC check retry times time out T1 T2 T3 T4 T5 T6 T7

Text information Contents menu rack no. /cup position rack no. digit
[PARAMETER] [ONLINE] [PROTOCOL] menu.
[PARAMETER] [FORMAT] [REQUISITION FORMAT] menu. patient information 1 patient information 2 patient information 3 patient information 4 patient information 5 patient information 6
use/none digits use/none digits use/none digits use/none digits use/none digits use/none digits
[PARAMETER] [FORMAT] [REQUISITION FORMAT] menu.
sample ID digits 4-26 digits [PARAMETER] [FORMAT] [REQUISITION FORMAT] menu. data format 6 /9 digits [PARAMETER] [ONLINE] [PROTOCOL] menu. data zero suppress used/none [PARAMETER] [ONLINE] [PROTOCOL] menu. Sample Type used/none [PARAMETER] [ONLINE] [PROTOCOL] [SPECIAL] menu. Dummy (Test Result Transmission Text)
used/none [PARAMETER] [ONLINE] [PROTOCOL] [SPECIAL] menu. Block Identification No. used/none [PARAMETER] [ONLINE] [PROTOCOL] [SPECIAL] menu. ANL Unit No. used/none Note 3 [PARAMETER] [ONLINE] [PROTOCOL] [SPECIAL] menu. Cuvette Side used/none Note 3 [PARAMETER] [ONLINE] [PROTOCOL] [SPECIAL] menu. Sequence of S.No. and S.ID in T.R.I text
S.No. / S.ID [PARAMETER] [ONLINE] [PROTOCOL] [SPECIAL] menu. Sequence of S.No. and S.ID in Test result text
S.No. / S.ID [PARAMETER] [ONLINE] [PROTOCOL] [SPECIAL] menu. Lot No. and Bottle No. used/none [PARAMETER] [ONLINE] [PROTOCOL] [SPECIAL] menu. Note 1)When you send the “Block Identification No.” field as none, the program will change the parameters automatically as following. End Code = ETX ETB Control = used Note 3) Only a Normal sample and repeat sample can be changed.
use/none 4/5 · Sex age/month · use/none use/none

Each of the following parameters are influenced as shown below.

RB R_ RH RE S_ SH SE DB D_ d_ DH dH DR DA DQ DE Sample Type - ○ × - ○ × - - ○ ○ × - Dummy (Test Result Transmission Text)
- - × - ○ × - - ○ ○ × -
- ○ × - × × - - ○ ○ × -
- - × - - × - × -
Sequence of S.No. and S.ID in Test result text
- ○ ○ × -
ANL Unit No. - ○ × × × - Cuvette Side - ○ × × × - Lot No. and Bottle No.
- ○ -


HIGH LOW 0 1 2 3 4 5 6 7 8 9 A B C D E F 0 DE SP 0 @ P p 1 SH D1 ! 1 A Q a q 2 SX D2 " 2 B R b r 3 EX D3 # 3 C S c s 4 ET D4 $ 4 D T d t 5 EQ NK % 5 E U e u 6 AK SN & 6 F V f v 7 BL EB ' 7 G W g w 8 BS CN ( 8 H X h x 9 HT EM ) 9 I Y i y A LF SB * : J Z j z B H M
C CL FS , < L \ l | D CR GS - = M ] m } E SO RS . > N ^ n ~ F SI US / ? O _ o DL
From 20H to 7EH: These characters can be used for the Sample ID code. From 30H to 39H: These characters can be used for Test numbers. A1 - DC is Japanese character set.
(3) Character Table
![Page 50 of the 2Wire AU2700 user manual — T2 [ ] (0-99)](/cdn/reader/8b/8bcde0ab16b25d439aec869989da7f7071970f3bf1e56c6bd9907bb6c3e1d6ff/images/page-50.webp?w=400)
(4) AU2700/5400 Online Parameter Sheet
Set up Setting T.R.I. Receive
Routine Normal real time batch none Routine Emergency real time batch none
Routine Repeat real time batch none
Emergency Repeat real time batch none Stat Normal real time batch none Stat Repeat real time batch none
Routine Normal real time batch none Routine Emergency real time batch none
Routine Repeat real time batch none
Emergency Repeat real time batch none
Stat Normal real time batch none
Stat Repeat real time batch none Stat Fast real time none
QC real time batch none Calibration real time batch none
Reagent blank real time batch none
Upper protocol Setting Error control
T.R.I. Receive continue stop Results Transfer continue stop
Lower protocol Setting Character format
Character length. 7 8 Parity bit odd even none
Stop bit 1 2
Start code 1 [ ] (01-1F) Start code 2 [ ] (00-1F)
End code 1 [ ] (01-1F)
End code 2 [ ] (00-1F)
Text length 256 512 1024
Unit No. [ ] (00-99 space)
ETB control yes no
SEQ. Control
Bit/sec 4800 9600 Class Class A Class B
BCC check yes no
Retry [ ] (0-3)
Time out T1 [ ] (0-99)
T3 [ ] (0-99)
T4 [ ] (0-99)
T5 [ ] (0-99)
T6 [ ] (0-99)
T7 [ ] (0-99)
Results Transfer · Data format · Text format Setting

Rack No. none 4 5 Sex yes no AGE/MONTH yes no
patient inf. 1 yes[ ]digits no
patient inf. 2 yes[ ]digits no
patient inf. 3 yes[ ]digits no
patient inf. 4 yes[ ]digits no
patient inf. 5 yes[ ]digits no
patient inf. 6 yes[ ]digits no
ID digits [ ]digits (4-26) Data format 6 9 ZERO Suppress yes no Sample Type yes no Dummy yes no Block Identification yes no ANL unit No. yes no Cuvette side yes no Sequence of S.No. and S.ID in T.R.I text
No. S.ID
Sequence of S.No. and S.ID in Test result text
yes no
Lot No. and Bottle No. yes no

1)ONLINE ERROR [ aa ] [ bbbbbbbbb ] (1). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive has been designated as "STOP", then the following will occur: 1) Text following the alarm will not be received. 2) In a case when the text requisition is received real-time,the "STOP" status is ignored at the start of the next analysis and the text requisition will be received real-time again. (2). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive has been designated to "CONTINUE", the test requisition for the next sample will be received real-time. (1).Communication error occurs when transmittng the text for test requisition or when receiving the test requisition format. (2).The display in the [ ] and its meaning are as follows. aa : type of the errors 01 : Device name error 02 : Framing error 03 : Over run error 04 : Parity error 05 : Time out error 06 : Receive NAK when transmitting a text 07 : BCC error when receiving a text 08 : Other communication errors 09 : Function error (Full Duplex instead of Half Duplex) 10 : Unit name error 11 : Parameter error 12 : Request cancel bbbbbbbbb : Text having error RB : Start text for receiving the test requisition information. RE : End text for receiving the test requisition information. R_ __ nnnn : normal run, normal sample, sample No. R_ _E nnnn : normal run, emergency sample, sample No. R_ _P nnnn : normal run, STAT sample, sample No. R_ U_ nnnn : normal run, urine sample, sample No. R_ UE nnnn : normal run, urine emergency sample, sample No. R_ UP nnnn : normal run, urine STAT sample, sample No. R_ X_ nnnn : normal run, urine sample, sample No. R_ XE nnnn : normal run, urine emergency sample, sample No. R_ XP nnnn : normal run, urine STAT sample, sample No. RH __ nnnn : repeat run, normal sample, sample No. RH _E nnnn : repeat run, emergency sample, sample No. RH _P nnnn : repeat run, STAT sample, sample No. RH U_ nnnn : repeat run, urine sample, sample No. RH UE nnnn : repeat run, urine emergency sample, sample No. RH UP nnnn : repeat run, urine STAT sample, sample No. R_ __ xxxxxxxxxxx: normal run, normal sample, sample ID R_ _E xxxxxxxxxxx: normal run, emergency sample, sample ID R_ _P xxxxxxxxxxx: normal run, STAT sample, sample ID R_ U_ xxxxxxxxxxx: normal run, urine sample, sample ID R_ UE xxxxxxxxxxx: normal run, urine emergency sample,sample ID R_ UP xxxxxxxxxxx: normal run, urine STAT sample, sample ID R_ X_ xxxxxxxxxxx: normal run, urine sample, sample ID R_ XE xxxxxxxxxxx: normal run, urine emergency sample,sample ID R_ XP xxxxxxxxxxx: normal run, urine STAT sample, sample ID RH __ xxxxxxxxxxx: repeat run, emergency sample, sample ID RH _E xxxxxxxxxxx: repeat run, emergency sample, sample ID RH _P xxxxxxxxxxx: repeat run, STAT sample, sample ID

RH U_ xxxxxxxxxxx: repeat run, urine sample, sample ID RH UE xxxxxxxxxxx: repeat run, urine emergency sample, sample ID RH UP xxxxxxxxxxx: repeat run, urine STAT sample, sample ID RH X_ xxxxxxxxxxx: repeat run, emergency sample, sample ID RH XE xxxxxxxxxxx: repeat run, emergency sample, sample ID RH XP xxxxxxxxxxx: repeat run, STAT sample, sample ID

2)ONLINE ERROR [ aa ] [ bbbbbbbbb ] (1). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive is designated as "STOP", the following events can occur: 1). The text following the alarm will not be received. 2). In a case when the text requisition is received real-time,the "STOP" designation is ignored and the text requisition will be received in real-time at the start of the next analysis. (2). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive is designated as "Continue", the test requisition for the next sample will be received. (1). Communication error occurs when transmitting the text for test requisition or receiving the test requisition format. (2). The display in the [ ] and its meaning are as follows. aa : type of the errors 01 : Device name error 02 : Framing error 03 : Over run error 04 : Parity error 05 : Time out error 06 : Receive NAK when transmitting a text 07 : BCC error when receiving a text 08 : Other communication errors 09 : Function error (Full Duplex instead of Half Duplex) 10 : Unit name error 11 : Parameter error 12 : Request cancel bbbbbbbbb : Text having error RB : Start text for receiving the test requisition information. RE : End text for receiving the test requisition information. R_ __ nnnn : normal run, normal sample, sample No. R_ _E nnnn : normal run, emergency sample, sample No. R_ _P nnnn : normal run, STAT sample, sample No. R_ U_ nnnn : normal run, urine sample, sample No. R_ UE nnnn : normal run, urine emergency sample, sample No. R_ UP nnnn : normal run, urine STAT sample, sample No. R_ X_ nnnn : normal run, Other normal sample, sample No. R_ XE nnnn : normal run, Other emergency sample, sample No. R_ XP nnnn : normal run, Other STAT sample, sample No. RH __ nnnn : repeat run, normal sample, sample No. RH _E nnnn : repeat run, emergency sample, sample No. RH _P nnnn : repeat run, STAT sample, sample No. RH U_ nnnn : repeat run, urine sample, sample No. RH UE nnnn : repeat run, urine emergency sample, sample No. RH UP nnnn : repeat run, urine STAT sample, sample No. RH X_ nnnn : repeat run, Other normal sample, sample No. RH XE nnnn : repeat run, Other emergency sample, sample No. RH XP nnnn : repeat run, Other STAT sample, sample No. R_ __ xxxxxxxxxxx: normal run, normal sample, sample ID R_ _E xxxxxxxxxxx: normal run, emergency sample, sample ID R_ _P xxxxxxxxxxx: normal run, STAT sample, sample ID R_ U_ xxxxxxxxxxx: normal run, urine sample, sample ID R_ UE xxxxxxxxxxx: normal run, urine emergency sample, sample ID R_ UP xxxxxxxxxxx: normal run, urine STAT sample, sample ID R_ X_ xxxxxxxxxxx: normal run, Other normal sample, sample ID R_ XE xxxxxxxxxxx: normal run, Other emergency sample, sample ID R_ XP xxxxxxxxxxx: normal run, Other STAT sample, sample ID RH __ xxxxxxxxxxx: repeat run, normal sample, sample ID RH _E xxxxxxxxxxx: repeat run, emergency sample, sample ID

RH _P xxxxxxxxxxx: repeat run, STAT sample, sample ID RH U_ xxxxxxxxxxx: repeat run, urine sample, sample ID RH UE xxxxxxxxxxx: repeat run, urine emergency sample, sample ID RH UP xxxxxxxxxxx: repeat run, urine STAT sample, sample ID RH X_ xxxxxxxxxxx: repeat run, Other normal sample, sample ID RH XE xxxxxxxxxxx: repeat run, Other emergency sample, sample ID RH XP xxxxxxxxxxx: repeat run, Other STAT sample, sample ID

3)ONLINE FORMAT ERROR (1).Disregard the test requisition information text for the sample generating the alarm. (2).Continue to receive the test requisition information text for samples following the alarm. (1).In the Online Parameter menu, submenu Set Up, the T.R.I. receive is not designated asn "YES" for the necessary parameter. (2).The display in the [ ] and its meaning are as follows. aaaa : Type of error R_ __ : normal run, a normal sample R_ _E : normal run, an emergency sample R_ _P : normal run, a STAT sample R_ U_ : normal run, a urine normal sample R_ UE : normal run, a urine emergency sample R_ UP : normal run, a urine STAT sample R_ X_ : normal run, a Other normal sample R_ XE : normal run, an Other emergency sample R_ XP : normal run, a Other STAT sample RH __ : repeat run, a normal sample RH _E : repeat run, an emergency sample RH _P : repeat run, a STAT sample RH U_ : repeat run, a urine normal sample RH UE : repeat run, a urine emergency sample RH UP : repeat run, a urine STAT sample RH X_ : repeat run, a Other normal sample RH XE : repeat run, an Other emergency sample RH XP : repeat run, a Other STAT sample bbbbbbbbbb : Information type that is not designated No sample No. No ID No.

4)ONLINE ILLEGAL TEXT CODE [ aa ] (1).Disregard the test requisition information text for the sample generating the alarm. (2).Continue to receive the test requisition information text for samples following the alarm. (1).When receiving the test requisition information text using online, the text classification code is out of specification. (2).The display of [ ] and its meaning are as follows. aa : R_ : Normal sample RH : Repeat run sample 5)ONLINE ILLEGAL TEXT BLOCK No. [ aa -> bb ] (1). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive is designated as "STOP", the following events occur: 1).Disregard the test requisition information text for the sample generating the alarm. 2).System will not receive text for samples following the sample generating the alarm. 3).In a case when the text requisition is received in real-time, the "STOP" is ignored and the text requisition text will be received with real-time again during the next analysis. (2). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive is designated as "Continue", the following events occur: 1).Disregard the test requisition information text for the sample generating the alarm. 2).The test requisition information text for the next sample will be received. (1).When receiving the requisition information text using online, the block identification No. is out of specification. (2).The display of [ ] and its meaning are as follows. aa : Previously received block identification No. bb : Last received block identification No.

6)ONLINE ILLEGAL SAMPLE No. [ aa: bbbbb ] (1).Disregard the test requisition information text for the sample generating the alarm. (2).Continue to receive the test requisition information text for other samples within the run. (1).When receiving the test requisition information text using online, sample No. of the sample generating the alarm does not meet the specification. (2).The display of [ ] and its meaning are as follows. aa : Sample type __ : Normal sample _E : Emergency sample _P : Stat sample U_ : Urine normal sample UE : Urine emergency sample UP : Urine stat sample X_ : Other Normal sample XE : Other Emergency sample XP : Other Stat sample bbbb : Sample No. 0001-9999 : Normal sample E001-E999 : Emergency P001-P999 : Stat sample

7)ONLINE ILLEGAL RACK No. [ aaaaaaa: b: cccccccc ] (1). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive is designated as "STOP", the following events occur: 1).Disregard the test requisition information text for the sample generating the alarm. 2).System will not receive text for samples following the sample generating the alarm. 3).In a case when the text requisition is received in real-time, the "STOP" is ignored and the text requisition text will be received with real-time again during the next analysis. (2). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive is designated as "Continue", the following events occur: 1).Disregard the test requisition information text for the sample generating the alarm. 2).The test requisition information text for the next sample will be received. (1).When receiving the requisition information text using online, the rack No. is out of specification. (2).The display of [ ] and its meaning are as follows. aaa : Transferred sample type and sample No. __ nnnn : Normal sample _E nnnn : Emergency sample _P nnnn : Stat sample U_ nnnn : Urine normal sample UE nnnn : Urine emergency sample UP nnnn : Urine stat sample X_ nnnn : Other normal sample XE nnnn : Other emergency sample XP nnnn : Other Stat sample bbb : Received sample type and sample No. __ nnnn : Normal sample _E nnnn : Emergency sample _P nnnn : Stat sample U_ nnnn : Urine normal sample UE nnnn : Urine emergency sample UP nnnn : Urine stat sample X_ nnnn : Other normal sample XE nnnn : Other emergency sample XP nnnn : Other Stat sample b : Received the kind of rack N : Normal sample rack E : Emergency rack R : Repeat run rack ccccccc : Received rack No. and position in rack

8)ONLINE ILLEGAL SAMPLE KIND No. [ aa ] (1).Disregard the test requisition information text for the sample generating the alarm. (2).Continue to receive the test requisition information text for the other samples within the run. (1).When receiving the test requisition information text using online, sample kind No. of the sample generating the alarm does not meet the specification. (2).The display of [ ] and its meaning are as follows. aa : Sample type __ : Normal sample _E : Emergency sample _P : Stat sample U_ : Urine normal sample UE : Urine emergency sample UP : Urine stat sample X_ : Other normal sample XE : Other emergency sample XP : Other stat sample 9)ONLINE ILLEGAL GENDER TEXT [ a ] (1).Disregard the test requisition information text for the sample generating the alarm. (2).Continue to receive the test requisition information text for other samples within the run. (1).When receiving the requisition information text using online, the sex of this text is not within the specification. (2).The display of [ ] and its meaning are as follows. a : Received sex 10)ONLINE ILLEGAL AGE/MONTH [ aaa bb ] (1).Disregard the test requisition information text for the sample generating the alarm. (2).Continue to receive the test requisition information text for other samples within the run. (1).When receiving the requisition information text using online, age or month age (age for baby before one year old) of this text is out of the specification. (2).The display of [ ] and its meaning are as follows. aaa : Received age bb : Received month age

11)ONLINE ANALYSIS METHOD MISMATCH [ aa bb ] (1).Disregard the test requisition information text for the sample generating the alarm. (2).Continue to receive the test requisition information text for other samples within the run. (1).When receiving the test requisition information text using online, sample type of this text does not match the required sample type. (2).The display of [ ] and its meaning are as follows. aa : Sample type __ : Normal sample _E : Emergency sample _P : Stat sample U_ : Urine normal sample UE : Urine emergency sample UP : Urine stat sample X_ : Other normal sample XE : Other emergency sample XP : Other Stat sample bb : Received sample type __ : Normal sample _E : Emergency sample _P : Stat sample U_ : Urine normal sample UE : Urine emergency sample UP : Urine stat sample X_ : Other normal sample XE : Other emergency sample XP : Other Stat sample
12)ONLINE SAMPLE No. MISMATCH [ aaaaaa bbbbbb ]

(1).Disregard the test requisition information text for the sample generating the alarm. (2).Continue to receive the test requisition information text for the other samples within the run. (1).When receiving the test requisition information text using online, sample type of this does not match the required sample. (2).The display of [ ] and its meaning are as follows. aaaaaaa : Transferred sample type and sample No. __ nnnn : Normal sample _E nnnn : Emergency sample _P nnnn : Stat sample U_ nnnn : Urine normal sample UE nnnn : Urine emergency sample UP nnnn : Urine stat sample X_ nnnn : Other normal sample XE nnnn : Other emergency sample XP nnnn : Other Stat sample bbbbbb : Received sample type and sample No. __ nnnn : Normal sample _E nnnn : Emergency sample _P nnnn : Stat sample U_ nnnn : Urine normal sample UE nnnn : Urine emergency sample UP nnnn : Urine stat sample X_ nnnn : Other normal sample XE nnnn : Other emergency sample XP nnnn : Other Stat sample

13)ONLINE RACK No. MISMATCH [ aaaaaa : bbbb cccc ] (1). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive is designated as "STOP", the following events occur: 1).Disregard the test requisition information text for the sample generating the alarm. 2).System will not receive text for samples following the sample generating the alarm. 3).In a case when the text requisition is received in real-time, the "STOP" is ignored and the text requisition text will be received with real-time again during the next analysis. (2). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive is designated as "Continue", the following events occur: 1).Disregard the test requisition information text for the sample generating the alarm. 2).The test requisition information text for the next sample will be received. (1).When receiving the test requisition information text using online, rack No. of this does not match the required sample. (2).The display of [ ] and its meaning are as follows. aaaaaaaaa : Transferred sample type and sample No. __ nnnn : Normal sample _E nnnn : Emergency sample _P nnnn : Stat sample U_ nnnn : Urine normal sample UE nnnn : Urine emergency sample UP nnnn : Urine stat sample X_ nnnn : Other normal sample XE nnnn : Other emergency sample XP nnnn : Other Stat sample bbbb : Transferred rack No. cccc : Received rack No. 14)ONLINE MISMATCH [ aaaaaaaaaa bbbbbbbbbbb ] (1).Disregard the test requisition information text for the sample generating the alarm. (2).Continue to receive the test requisition information text for the other samples within the run. (1).When receiving the test requisition information text using online, sample ID of this text does not match the required ones. (2).The display of [ ] and its meaning as follows. aaaaaaaaa : Transferred sample ID bbbbbbbbb : Received sample ID

15)ONLINE ITEM ERROR [ aaaaaaaa ] (1).Disregard the test requisition information text for the sample generating the alarm. (2).Continue to receive the test requisition information text for the other samples within the run. (1).When receiving the test requisition information text using online, the test requisition cannot be stored. (2).The display of [ ] and its meaning are as follows. aaaaaa : Sample type and sample No. __nnnn : Normal sample __Ennn : Emergency sample __Pnnn : Stat sample _U_nnn : Urine normal sample _UEnnn : Urine emergency sample _UPnnn : Urine stat sample _Xnnnn : Other normal sample _XEnnn : Other emergency sample _XPnnn : Other Stat sample H_nnnn : repeat run normal sample H_Ennn : repeat run emergency sample H_Pnnn : repeat run stat sample HUnnnn : repeat run urine normal sample HUEnnn : repeat run urine emergency sample HUPnnn : repeat run urine stat sample HXnnnn : repeat run other normal sample HXEnnn : repeat run other emergency sample HXPnnn : repeat run other stat sample

16)ONLINE RERUN ITEM ERROR [ aaaaaaaa ] (1).Disregard the repeat run test requisition information for the sample generating the alarm. (2).Continue to receive the test requisition information text for the other samples within the run. (1).When receiving the test requisition information text for repeat run using online, repeat run test requisition cannot be stored with for the following reasons. 1).Original sample that is designated in the repeat run text is not stored in the normal run/sample test requisition information. 2).Sample No. is not set to the original sample that is designated in the repeat run text. 3).Original sample that is designated in the repeat run text has already been stored as a repeat run sample for another sample. (2).The display of [ ] and its meaning are as follows. aaaaaa : sample type and original sample No. __nnnn : Normal sample __Ennn : Emergency sample __Pnnn : Stat sample _Unnnn : Urine normal sample _UEnnn : Urine emergency sample _UPnnn : Urine stat sample _Xnnnn : Other normal sample _XEnnn : Other emergency sample _XPnnn : Other Stat sample
Owners and our team usually reply within a day — and your question helps the next person too.