Page 1

OLYMPUS DIAGNOSTIC SYSTEMS GROUP

AU2700 & AU5400 ON-LINE SPECIFICATIONS

Page 2 of the 2Wire AU2700 user manual — AU2700/5400
Page 2 · click to enlarge
Page 2

AU2700/5400

Online Specifications

Issued2000.3.16 Revised: 2000.6.8

Olympus Optical Co., Ltd.

Page 4 of the 2Wire AU2700 user manual — 2. Basic Specification
Page 4 · click to enlarge
Page 4

1. Outline

1. The following data is exchanged between the AU2700/5400 and the host computer system.

(1) Test Requisition Information (T.R.I.) (2) Test Results

2. The data transmission is conducted in either of the following mode:

(1) Class A Transmission is without synchronizing with analysis. This mode does not utilize the ACK/NAK protocol. (2) Class B Transmission is synchronizing with analysis. This mode does utilize the ACK/NAK protocol.

3. The Following format and protocol can be selected by [PARAMETER] [ONLINE] menu.

(1) Setup (2) Communication protocol

A. T.R.I. Inquiry Text Format B. Test Result Text Format (3) Test Number Assignment

2. Basic Specification

Page 5 of the 2Wire AU2700 user manual — 2. Basic Specification
Page 5 · click to enlarge
Page 5

(2.1) Format of Transmission

2Wire AU2700 — 2. Basic Specification diagram (page 5)
Fig. 1 · 2. Basic Specification
ItemContents Transmission lineRS-232C SynchronizationSynchronization Data transmission modeHalf-Duplex Bit/sec4800bps, 9600bps Configurationstart bit 1 bit data bit 7 or 8 bit parity bit none, odd or even stop bit 1 or 2 bit total 9-12 bit These can be selected by [PARAMETER] [ONLINE] [PROTOCOL] menu. Form of TransmissionClass A Transmission are conducted from the sender to the receiver in a certain time of interval. Class B Transmission are conducted by confirming with ACK (receipt acknowledged)or NAK (not acknowledged) between the sender and the receiver. Use channel1 channel RetryClass A: None Class B: 0-3 This can be selected by [PARAMETER] [ONLINE] [PROTOCOL]menu.
start bit1 bit data bit7 or 8 bit parity bitnone, odd or even stop bit1 or 2 bit total9-12 bit
2Wire AU2700 — 2. Basic Specification diagram (page 5)
Fig. 2 · 2. Basic Specification
ItemContentsLimit of value Data code These can be selected by [PARAMETER] [ONLINE] [PROTOCOL] menu.7 bit code ASCII CODE20H-7EH 8 bit code 1 byte code 1 byte code20H-7EH A1H-DFH Control codeStart code/terminate code01H-1FH ACK06H NAK15H BCC00H-FFH
Page 6 of the 2Wire AU2700 user manual — Text format
Page 6 · click to enlarge
Page 6

(2.3) Text format (1) Basic configuration as follows:

1. 2) 3) 4) 5) 6)
NameDigitContentsRemarks 1) Text start code1,201h-1FhData start codeNormally 02h 2) Text Classification2R# RB R * _ RH RET.R.I. inquiryAU2700/5400 → Host RBStart of T.R.I. inquiry R * _T.R.I. inquiry of normal sample RHT.R.I. inquiry of rerun sample REEnd of T.R.I. inquiry S# S * _ SH SET.R.I.Host → AU2700/5400 S *T.R.I. of normal sample _ SHT.R.I. of rerun sample SEStop of T.R.I. inquiry D# DB D * _ DH DR DA d * _ dH DQ DETest resultAU2700/5400 → Host DBStart of transfer result D *Normal sample result _ DHRerun sample result DRReagent blank sample Result DACalibration result d * _Stat fast result(AU2700 Only) Emergency fast result (AU5400 Only) dHStat fast rerun result (AU2700 Only) Emergency fast rerun result (AU5400 Only) DQQC sample result DEEnd of result transfer 3) Unit No. (Optional)0,2[PARAMETER] [ONLINE] [PROTOCOL] menu. 4) Text codeContents of Text[PARAMETER] [ONLINE] [PROTOCOL] menu. Data No. is added behind text of Header parts. 5) text end code1,201h-1FhEnd of text code[PARAMETER] [ONLINE] [PROTOCOL] menu. 6) BCC (Block Check Character) (Optional)100h-FFhThe sum with the Exclusive or logic between 2-5 and including 5.[PARAMETER] [ONLINE] [PROTOCOL] menu.

remarks * _ is space.

Page 7 of the 2Wire AU2700 user manual — Text format
Page 7 · click to enlarge
Page 7

A. Definition of Term

1. 2) 3) 4) 5) 6)
TermDefinition Text length [PARAMETER] [ONLINE] [PROTOCOL] menu.Is the total number of bytes from 1) to 6). Max block length [PARAMETER] [ONLINE] [PROTOCOL] menu.It is the Max length of text in one record. The length of block can be selected in online parameter menu. (256, 512, 1024)

(2) Blocking

Page 8 of the 2Wire AU2700 user manual — Text format
Page 8 · click to enlarge
Page 8

B. Kind of text and blocking

KindKind of textMethod of Blocking Fixed length Text• Start of T.R.I. inquiry • T.R.I. inquiry text • End of T.R.I. transmission text. • Stop of T.R.I. inquiry • Start of transfer result transmission text. • End of result transfer transmission text.Non-blocking Variable length text• T.R.I. text • Rerun T.R.I. text • Test result text • Rerun result text • QC result text • Acal result text • Reagent blank result text • STAT(fast) result text (AU2700 only) • Emergency(fast) result text (AU5400 only)Using blockingMax block length < text length Block Identification No. (0-9,E) (see page 9)first block Block Identification No. =0 second block Block Identification No. =1 • • last block Block Identification No. =E Text end codeText end code = ETX(03H) and using ETB = yes *1 End of block = ETB End of text = ETX Text end code <>ETX(03H) or ETB(17h)is no-use *1 End of block is designated in menu. End of text is designated in menu. Using ETB can be selected in online parameters.

Remark *1:When Block Identification No. is numeric(0,1, ), Only “End of Block” is used.

When Block Identification No. is “E”, Only “End of Text” is used.

Page 9 of the 2Wire AU2700 user manual — 1) R B 3) 5) 6)
Page 9 · click to enlarge
Page 9

3. Transmission Text Format

(3.1) Common and changeable item

The following can be selected by [PARAMETER] [ONLINE] [PROTOCOL] menu. Text start code 1) Unit No. 3) Text end code 5) BCC 6)

Sample Type Dummy(in Test Result transmission Text) Block Identification No. Rack No./Cup Position Unit No. on the T.R.I. text (with out Rerun) ANL unit No. and Cuvette side on Test Result transmission Text (without RB,ACAL,QC) Space Code before the Text Classification (S,R,D text) Reagent Lot No. and bottle No.

The order of the Sample number and Sample ID on the text can be changed, the sample type, the Dummy spaces and the Block Identification Number can be changed in the [PARAMETER] [ONLINE] [PROTOCOL] menu. There is a function at the bottom of the page called “Special”. These features may be altered under the “Special” menu. Making changes in the “Special” menu only affects the T.R.I. Text and Test Result Transmission Text.

(3.2) T.R.I. inquiry text

(1) Start text to inquiry test requisition.

1) R B 3) 5) 6)

(2) Normal run inquiry test requisition.

2Wire AU2700 — 1) R B 3) 5) 6) diagram (page 9)
Fig. 3
1)R _3)Rack no.Cup PositionSample no.Samp

Common Item

Page 10 of the 2Wire AU2700 user manual — 1) R E 3) 5) 6)
Page 10 · click to enlarge
Page 10

(3) Repeat run inquiry test requisition.

2Wire AU2700 — 3. Transmission Text Format diagram (page 10)
Fig. 4 · 3. Transmission Text Format

(4) End text to inquiry test requisition.

1) R E 3) 5) 6)

Sample type

Page 11 of the 2Wire AU2700 user manual — T.R.I. text
Page 11 · click to enlarge
Page 11

(3.3) T.R.I. text

(1) Normal sample test requisition information text.

2Wire AU2700 — T.R.I. text diagram (page 11)
Fig. 5 · T.R.I. text
1)S _Rack No.cup positionSample no.Samp

Sample type Block Identification No.

Test code 5) 6)

(2) Repeat run sample test requisition information text.

2Wire AU2700 — T.R.I. text diagram (page 11)
Fig. 6 · T.R.I. text
1)S HRack no.Cup PositionSample no.Samp

Original Sample no.

Sample type Block Identification No.

Blocking Test Me code th od Me th odTest codeMe th odTest code tMe h odTest codeMe th odTest codeMe th od5)6)
2Wire AU2700 — T.R.I. text diagram (page 11)
Fig. 7 · T.R.I. text
Show 15 more labels / short text on this page
s e x · Patient Information 1 · Patient Information 2 · Patient Information 3 · age month · Patient Information 4 · Patient Information 5 · Patient Information 6 · Test code · Test code · Test code · Test code · Test code · Test code · Test code
Page 12 of the 2Wire AU2700 user manual — 1) D B 3) 5) 6)
Page 12 · click to enlarge
Page 12

(3) End of T.R.I. Text transmission.

1) S E 5) 6)

(3.4) Test result transmission text

(1) Start of test result transmission text.

1) D B 3) 5) 6)

Page 13 of the 2Wire AU2700 user manual — Test result transmission text
Page 13 · click to enlarge
Page 13

(2) Normal sample test result text.

2Wire AU2700 — Test result transmission text diagram (page 13)
Fig. 8 · Test result transmission text

Sample no. Sample ID Dummy

Sample type Block Identification No.

2Wire AU2700 — Test result transmission text diagram (page 13)
Fig. 9 · Test result transmission text
Show 7 more labels / short text on this page
S e x · Patient Information 1 · Patient Information 2 · Patient Information 3 · Age Month · Blocking · Blocking
Page 14 of the 2Wire AU2700 user manual — Test result transmission text
Page 14 · click to enlarge
Page 14

(3) Stat(fast) test result text (AU2700 only).

2Wire AU2700 — Test result transmission text diagram (page 14)
Fig. 10 · Test result transmission text

P n n n Sample no. Sample ID Dummy

1. d _ 3) Rack no

1-22 Sample type Block Identification No.

2Wire AU2700 — Test result transmission text diagram (page 14)
Fig. 11 · Test result transmission text
Show 7 more labels / short text on this page
S e x · Patient Information 1 · Patient Information 2 · Patient Information 3 · Age Month · Blocking · Blocking
Page 15 of the 2Wire AU2700 user manual — Test result transmission text
Page 15 · click to enlarge
Page 15

(4) Emergency(fast) test result text (AU2700 only).

2Wire AU2700 — Test result transmission text diagram (page 15)
Fig. 12 · Test result transmission text

E n n n Sample no. Sample ID Dummy

1. d _ 3) Rack no

1-22 Sample type Block Identification No.

2Wire AU2700 — Test result transmission text diagram (page 15)
Fig. 13 · Test result transmission text
Show 7 more labels / short text on this page
S e x · Patient Information 1 · Patient Information 2 · Patient Information 3 · Age Month · Blocking · Blocking
Page 16 of the 2Wire AU2700 user manual — Test result transmission text
Page 16 · click to enlarge
Page 16

(5) Repeat run result text -1 .

2Wire AU2700 — Test result transmission text diagram (page 16)
Fig. 14 · Test result transmission text

Original Sample no.

Sample type Block Identification No.

2Wire AU2700 — Test result transmission text diagram (page 16)
Fig. 15 · Test result transmission text

Blocking · Blocking

Page 17 of the 2Wire AU2700 user manual — Test result transmission text
Page 17 · click to enlarge
Page 17

(6) Repeat run result text -2 Stat fast sample (AU2700 Only) .

2Wire AU2700 — Test result transmission text diagram (page 17)
Fig. 16 · Test result transmission text

Original Sample No.

1. d H 3) Rack no

1-22 Sample type Block Identification No.

2Wire AU2700 — Test result transmission text diagram (page 17)
Fig. 17 · Test result transmission text

Blocking · Blocking

Page 18 of the 2Wire AU2700 user manual — Test result transmission text
Page 18 · click to enlarge
Page 18

(7) Reagent blank result text.

2Wire AU2700 — Test result transmission text diagram (page 18)
Fig. 18 · Test result transmission text
1. D R 3) Sample no

This dummy is same ID digits Block Identification No. and assigned area is decided associated with the rack No. ANL unit No. Cuvette side

Test code R1 Lot No. R1 Bottle No. R2 Lot No. R2 Bottle No.

Page 19 of the 2Wire AU2700 user manual — Test result transmission text
Page 19 · click to enlarge
Page 19

(8) Acal (Calibration) result text.

2Wire AU2700 — Test result transmission text diagram (page 19)
Fig. 19 · Test result transmission text
1. D A 3) Sample no. Calibrator ID

This dummy spaces is assigned in case rack No. exists Calibrator No. (1-80) and assigned area is decided associated with the rack No. Block Identification No.

2Wire AU2700 — Test result transmission text diagram (page 19)
Fig. 20 · Test result transmission text

Blocking · Blocking

Page 20 of the 2Wire AU2700 user manual — 1) D E 3) 5) 6)
Page 20 · click to enlarge
Page 20

(9) QC (Quality Control) result text.

2Wire AU2700 — Test result transmission text diagram (page 20)
Fig. 21 · Test result transmission text
1. D Q 3) Sample no. Control ID

This dummy No. is assigned in case rack No. exists Control No. and assigned area is decided associated with the rack No. Block Identification No. ANL unit No.

2Wire AU2700 — Test result transmission text diagram (page 20)
Fig. 22 · Test result transmission text

(10) End of test result transfer text.

1) D E 3) 5) 6)

Blocking · Blocking

Page 21 of the 2Wire AU2700 user manual — Relation between real-time/batch and Transmission text
Page 21 · click to enlarge
Page 21

(3.5) Relation between real-time/batch and Transmission text

  • T.R.I. inquiry of Routine, Emergency, Stat and Repeat run can be selected as real-time or batch.
  • Test results of all samples can be selected as real-time or batch.
Page 22 of the 2Wire AU2700 user manual — Relation between real-time/batch and Transmission text
Page 22 · click to enlarge
Page 22

(3.6) Others

(1) Contents and formats of text.

Item Digit Contents Remarks 1. Rack No.

4 or 5 '0001'-'9999' or '00001'-'99999' STAT sample is space

  • The digit of rack no. and cup pos. and these using is changeable in [PARAMETER] [ONLINE] [PROTOCOL] menu .

When Rack no. is no setting, Rack No. is '00000'.

  • In case ,receive T.R.I. in batch. 2. Cup position (Optional)
  • This is automatically assigned when rack No. is in use. 3. Sample type 1 space : serum U : urine X : other 4. Sample No. 4 '0001'-'9999' : Routine sample 'E001'-'E999' : Emergency sample 'P001'-'P999' : Stat sample 'R001'-'R999' : Reagent blank 'A001'-'A999' : Calibration sample 'Q001'-'Q999' : QC sample 5. Sample ID 4-26 Number or character
  • Number of digits can be changed in the [PARAMETER] [FORMAT] [FORMAT REQUISITION] menu. 6. Original Sample No.
2 rack'01'-'10' STAT : '01'-'22'
4 '0001'-'9999'normal sample 'E001'-'E999' : emergency sample 'P001'-'P999' : stat sample

For Repeat Run Results.

7. Reserved 4 Space 8. Block 1 '0'-'9' Last text is 'E' Identification No.

(Optional)

Page 23 of the 2Wire AU2700 user manual — ‘1’: AU2700
Page 23 · click to enlarge
Page 23

9.ANL unit No.

ANL unit No. is added to the header portion of RB, ACAL, and QC result data. Normal sample and rerun sample, it is added to each item. However, in the case of an calculate item, a space is added. In a normal sample and repeat run, it can be prevented from adding a unit number by the setup of an on-line condition menu. This option can be selected in the [PARAMETER][ONLINE] [PROTOCOL][SPECIAL] menu. 10.Cuvette side

1 ‘1’AU5400 ISE ‘1’: AU5400 unit 1 ‘2’: AU5400 unit 2 ‘3’: AU5400 unit 3

‘1’: AU2700

Cuvette side is added to the header portion of RB, ACAL, and QC result data. Normal sample and rerun sample, it is added to each item. However, in the case of an calculate item, a space is added. In a normal sample and rerun, it can be prevented from adding a unit number by the setup of an on-line condition menu. This option can be selected in the [PARAMETER][ONLINE] [PROTOCOL][SPECIAL] menu. 11. Sex

1 AU5400 ‘0’ : inner cuvette (ISE Cell 1) ‘1’ : outer cuvette (ISE Cell 2) AU2700 ‘0’ : inner cuvette(ISE) ‘1’ : outer cuvette

  • These items can be selected/canceled in the [PARAMETER] [FORMAT] [FORMAT REQUISITION] menu. 12. Age (Optional)
1 Mmale F : female SP : none sex 0 : no-set
3 '000'-'150' spaceno-set *1 13. Month (Optional)
2 '00'-'11' spaceno-set *1 14. Patient Information (Optional)

20 number or character

15. Online Test No.

2 '01'-'99' • This number can be changed in the [PARAMETER] [ONLINE] [TEST NO.] 16. Method 1 '0':normal '1':dilution ‘2’:condense

For Repeat Run Results.

(Optional) · (Optional) · (Optional)

Page 24 of the 2Wire AU2700 user manual — Relation between real-time/batch and Transmission text
Page 24 · click to enlarge
Page 24
17. Test result 8 or 11 Test Result digits can be 6 or 9. Flags digits is 2.
  • 0 suppress can be selected/canceled in online parameter menu

Refer to Appendix 1 (Flags List)

  • Number of digits can be changed to 6 or 9 in online parameter menu. 18. Lot No. and Bottle No. (Optional)

4*4 Number or character • This option can be selected in the [PARAMETER] [ONLINE] [PROTOCOL] [SPECIAL] menu.

Remark *1: If only age is the space in the text from a host, AU2700/AU5400 will output an error. The combination of Age

and Month is as follows.

Pattern Age Month Result 1 Spaces Spaces OK 2 XXX Spaces OK 3 Spaces XX Error 4 XXX XX OK

Host -> AU2700/5400

Page 25 of the 2Wire AU2700 user manual — Relation between real-time/batch and Transmission text
Page 25 · click to enlarge
Page 25

(2) Test result format.

A. Test result

  • Data digit is 6 and 0 suppress is not used.

0 1 2 3 . 4

  • Data digit is 9 and 0 suppress is not used.

- 0 1 2 3 . 4 5 6

  • Data digit is 6 and 0 suppress is used.

- _ 1 2 3 4

  • Data digit is 9 and 0 suppress is used.

- _ 1 2 3 . 4 5 6

Page 26 of the 2Wire AU2700 user manual — LIP ICT HEM
Page 26 · click to enlarge
Page 26

B. LIH result

  • Data digit is 6 and 0 suppress is not used.

LIP ICT HEM

0 NORMAL FLAG 1 + 2 ++ 3 +++ 4 ++++ 5 +++++ 6 ABN 9 Test not performed.

  • Data digit is 9 and 0 suppress is not used.

LIP ICT HEM

0

0 NORMAL FLAG 1 + 2 ++ 3 +++ 4 ++++ 5 +++++ 6 ABN 9 Test not performed.

Page 27 of the 2Wire AU2700 user manual — LIP ICT HEM
Page 27 · click to enlarge
Page 27
  • Data digit is 6 and 0 suppress is used.

LIP ICT HEM

0 NORMAL FLAG 1 + 2 ++ 3 +++ 4 ++++ 5 +++++ 6 ABN 9 Test not performed.

  • Data digit is 6 and 0 suppress is used.

LIP ICT HEM

0 NORMAL FLAG 1 + 2 ++ 3 +++ 4 ++++ 5 +++++ 6 ABN 9 Test not performed.

_ _ _

Page 28 of the 2Wire AU2700 user manual — Relation between real-time/batch and Transmission text
Page 28 · click to enlarge
Page 28

(3) T.R.I. of calculated tests.

(4) Data format that Test result data is digit over.

a. Data digit is 6.

9

  • Data digit is 9.

9

(5) Data format that Test result data is OD-value.

  • Data digit is 6.
0. 1 2 3 4

When the data is negative OD-value, it is same as digit over.

  • Data digit is 9.

- 0 . 1 2 3 4

Page 29 of the 2Wire AU2700 user manual — Relation between real-time/batch and Transmission text
Page 29 · click to enlarge
Page 29

(6) T.R.I of LIH.

In any case, LIH test no. in T.R.I from Host is necessary for LIH analysis.

LIH Reagent Test Name LIH Measure All Even if AU2700/AU5400 does not have the request from a host, it performs LIH analysis. LIH

Select AU2700/AU5400 performs LIH analysis according to the request from a host. All Even if AU2700/AU5400 does not have the request from a host, it performs LIH analysis. Other

No LIH analysis is not performed even if AU2700/AU5400 has the request from a host. Note 1) These can be selected by the following menus.

LIH Reagent Test Name : [PARAMETER] – [COMMON TEST PARAMETERS] – [TEST NAME] LIH Measure : [PARAMETER] – [COMMON TEST PARAMETERS] – [ROUND]

Page 30 of the 2Wire AU2700 user manual — 4. Transmission Protocols
Page 30 · click to enlarge
Page 30

4. Transmission Protocols

(4.1) Protocols at AU2700/5400

(1) Class A: without ACK/NAK exchange.

A. Sending

Text to be sent:

T.R.I. Inquiry text Repeat Run T.R.I. Inquiry text

Start of Test Result Transmission text Test Result text Repeat Run Result text End of Test Result Transmission text

Start Block end (BCC) Initial of Transmission text Condition Text Text end (BCC)

B. Receiving

Text to be received:

T.R.I. text

Repeat Run T.R.I. text End of T.R.I. Transmission text

Start Block end (BCC) Initial of Transmission text Condition Text Text end (BCC)

Page 31 of the 2Wire AU2700 user manual — 4. Transmission Protocols
Page 31 · click to enlarge
Page 31

(2) Class B: with ACK/NAK protocol.

A. Sending

Text to be sent:

T.R.I. Inquiry text Repeat Run T.R.I. Inquiry text Start of Test Result Transmission text Test Result text Repeat Run Result text End of Test Result Transmission text

ACK ACK Received Received

Block end (BCC) Initial Start Condition of Transmission text Text Text end (BCC)

NAK NAK Received Received 1st time: resending the same block or or 2nd time: sending the next block Time-out Time-out

1 st timeresending the same text 2 nd time: return to initial condition
Page 32 of the 2Wire AU2700 user manual — 4. Transmission Protocols
Page 32 · click to enlarge
Page 32

B. Receiving

Test to be sent:

T.R.I. text Repeat Run T.R.I. text End of T.R.I. Transmission text

ACK ACK Send Send

Block end (BCC) Initial Start Condition of Transmission text Text Text end (BCC)

NAK NAK Send Send 1st time: receiving the same block 2nd time: receiving the next block

1 st timereceiving the same text 2 nd time: return to initial condition
Page 33 of the 2Wire AU2700 user manual — A B C
Page 33 · click to enlarge
Page 33

(4.2) Timing/time-out

(1) Time-out/timing list .

Kind Description Default Remarks T1 Limit time from end of transfer/receive to start of text receive. 2 s 0.1 x n sec T2 Limit time from start of text receive to end of text receive. (Note 1)

T3 Limit time from start of text transfer to end of text transfer. n=1-99 T4 Limit time from end of text transfer to response of receive. 2 s T5 Interval time of transmission 2 s T6 Min. time from receive 1 s T7 Limit time from NAK response to start receive of retry text.

t1 Min. time from end of transfer text to receivable 0.5 s Not Changeable text. t2 Min. time from end of transfer text to next receivable text. 0.5 s

Min. time from end of transfer text to receivable Response. 0.5 s

Min. time from end of receiving text to transfer Response. 0.5 s

(2) Rule of class A

2Wire AU2700 — Timing/time-out diagram (page 33)
Fig. 23 · Timing/time-out

<T3 <T2 <T3

B. Case 2

A B C

AU2700/5400

<T3 <T3 <T3

T5< T5<

Show 6 more labels / short text on this page
(Note 2) · 2 s · t3 · t4 · t1<<T1 T5< · Host
Page 34 of the 2Wire AU2700 user manual — <T2 <T2 <T2
Page 34 · click to enlarge
Page 34

C. Case 3

AU2700/5400 Host A B C

<T2 <T2 <T2

t2

Page 35 of the 2Wire AU2700 user manual — <T3 <T3 <T3
Page 35 · click to enlarge
Page 35

(4.3) Rule of class B

2Wire AU2700 — Rule of class B diagram (page 35)
Fig. 24 · Rule of class B

<T3 <T2 <T3

t3

2Wire AU2700 — <T3 <T2 <T3 diagram (page 35)
Fig. 25

<T3 <T3

2Wire AU2700 — <T3 <T3 diagram (page 35)
Fig. 26

<T2 <T2

(4) Case 4 (Abnormal Case 1)

2Wire AU2700 — <T2 <T2 diagram (page 35)
Fig. 27

<T3 <T3 <T3

t3

Show 5 more labels / short text on this page
(1) Case 1 · (2) Case 2 · t3<<T4 T5< · (3) Case 3 · t4< t2<<T1
Page 36 of the 2Wire AU2700 user manual — <T2 <T2 <T3
Page 36 · click to enlarge
Page 36

(5) Case 5 (Abnormal Case 2)

2Wire AU2700 — Rule of class B diagram (page 36)
Fig. 28 · Rule of class B

<T3 <T3 <T2

t3

(6) Case 6 (Abnormal Case 3)

2Wire AU2700 — <T3 <T3 <T2 diagram (page 36)
Fig. 29

<T2 <T2 <T2

t4< t2

(7) Case 7 (Abnormal Case 4)

2Wire AU2700 — <T2 <T2 <T2 diagram (page 36)
Fig. 30

<T2 <T2 <T3

t4< t2

Page 37 of the 2Wire AU2700 user manual — 5. Application Protocol
Page 37 · click to enlarge
Page 37
5. Application Protocol (5.1) Receiving T.R.I.

(1) Trans./receive sequence in one session.

General sequence :

A. Example 1 Shift next Rx-transfer when Online Error occurred.

2Wire AU2700 — 5. Application Protocol diagram (page 37)
Fig. 31 · 5. Application Protocol

B. Example 2 Stop the session when Online Error occurred.

2Wire AU2700 — 5. Application Protocol diagram (page 37)
Fig. 32 · 5. Application Protocol

(2) Detail T.R.I. Receive Kind of text Timing/condition Real-time RB(request This text is transferred when it is shifted from stand-by Start) mode to start measure and then the first Rx is transmitted.

R_ * (request normal) (Note 2) Request of When detecting cup, this Sample No. is transferred if this sample no. sample has not T.R.I.. Request of When reading ID is normally, this Sample ID is transferred sample id if this sample has not T.R.I.. RH (request rerun) (Note 2) Request of When detecting cup, this Sample No. is always transmitted sample no. Request of When reading ID is normally, this Sample ID is always sample ID transmitted. remarks * _ is space. Note 1) The number of times a T.R.I. inquiry can be sent when an online error is generated can be selected in [PARAMETER] [ONLINE] [PROTOCOL]. Note 2) It is possible that the R_-text and RH-text are mixed in one session.

Page 38 of the 2Wire AU2700 user manual — 5. Application Protocol
Page 38 · click to enlarge
Page 38

Receive Kind of text Timing/condition

This text can be received after end of Rx-transfer within the designated time. SE RE (request This text is transferred when shifting to the modes below. end) 1) measure mode to standby mode 2) measure mode to stop mode

This text is transmitted when the communication is canceled by the online alarm. Batch RB (request This text is transmitted when receiving T.R.I. is sent in the test start) requisition.

R_ * (request Normal) (Note 3) Request of This text is transmitted at the designated time intervals. sample no. The texts are the samples that are selected in the test requisition menu.

RH (request rerun) (Note 3)

Request of This text is transmitted at the designated time intervals. sample no. The texts are the samples that are selected in the test requisition menu. Sx This text can be received after end of Rx-transfer within the designated time. SE

RE (request end) This text is transmitted when it was receiving the last sample No. in test requisition menu and SE text isn't received. This text is transmitted when the communication is canceled in the test requisition menu. This text is transmitted when the communication is canceled by the online alarm. remarks * _ is space. Note 3) R_-text and RH-text is transferred in separate sessions.

Real-time Sx

Page 39 of the 2Wire AU2700 user manual — 5. Application Protocol
Page 39 · click to enlarge
Page 39

(3) Kind of T.R.I. transfer/receive.

A. T.R.I. of normal sample

2Wire AU2700 — 5. Application Protocol diagram (page 39)
Fig. 33 · 5. Application Protocol

Setting of parameters Method of T.R.I. inquiry & T.R.I. Receive T.R.I.. inquiry Method of T.R.I. inquiry

Sample no (Calculated rack No. and cup pos. No.) Sample ID Request of sample ID

Sample ID *1 Sample no *1 Batch Sequential Rack no.

  • Selectable setting of receiving T.R.I. in online parameter menu and requisition format menu.
  • Selectable method of T.R.I. in system menu. Remark *1: Sample no. and Sample ID of T.R.I. must be same to the Sample no. and Sample ID of T.R.I inquiry.

(4) T.R.I. of Repeat sample

2Wire AU2700 — 5. Application Protocol diagram (page 39)
Fig. 34 · 5. Application Protocol

Setting of parameters Method of T.R.I. inquiry & T.R.I. Repeat sample Receive of T.R.I. Repeat rack

Original sample no. *1 Repeat sample no.*1 Sample ID Request of sample ID

Remark *1: Repeat sample no. and sample ID of T.R.I. must be same as the Repeat sample no. and sample ID of T.R.I

inquiry.

Page 40 of the 2Wire AU2700 user manual — Sending test result protocol
Page 40 · click to enlarge
Page 40

(5.2) Sending test result protocol

(1) Transfer sequence for one session.

General sequence :

A. Example 1

Shift next Dx-transfer when Online Error occurred.

2Wire AU2700 — Sending test result protocol diagram (page 40)
Fig. 35 · Sending test result protocol

B. Example 2

Stop session when Online Error occurred.

2Wire AU2700 — Sending test result protocol diagram (page 40)
Fig. 36 · Sending test result protocol

(2) Detail Data Trans. Kind of text Timing/condition Normal end Online error Real DB (START) This text is transmitted when It is shifting the It is output the -time it is shifted from stand-by transfer of the alarm and it is mode to measure mode and just test results. executed the below behind the first Dx is transmitted.

  • Error protocol is Dx (result data) It is transmitted one by one It is shifting next stop. the samples that are finished The transfer of It is canceled the measuring. Dx/DE. transfer of Dx/DE
  • Error protocol is continue It is shifting next the transfer of Dx/DE.

DE (end) This text is transmitted after It is ended the It is output the the last transfer of Dx text session of transfer of alarm and ended and the below shifting mode. The text of test the session of the results. transfer of the 1) measure mode to standby text of test results. 2) measure mode to stop This text is transmitted when the communication is canceled by the online alarm.

2 blocks Host · Host

Page 41 of the 2Wire AU2700 user manual — Sending test result protocol
Page 41 · click to enlarge
Page 41

Data Trans. Kind of text Timing/condition Normal end Online error Batch DB (start) This text is transmitted when it It is shifting the It is output the selected the start of transmission transfer of the test alarm and it is in the online menu. Results. executed the below

  • Error protocol is This text is transmitted at It is shifting next stop. Dx the designate time intervals. The transfer of It is canceled the (result data) The text are the samples that Dx/DE. transfer of Dx/DE is selected in the online menu
  • Error protocol is continue. It is shifting next the transfer of Dx/DE.

DE (end) This text is transmitted when It is ended the It is output the it was transmitting the session of transfer alarm and ended sample No. that is selected in of the text of test The session of the online menu and it was after results. transfer of the the designated time. text of test results.

This text is transmitted when the communication is canceled in the online menu. This text is transmitted when the communication is canceled by the online alarm.

[PARAMETER] [ONLINE] [PROTOCOL] menu.

Page 42 of the 2Wire AU2700 user manual — AU2700/5400
Page 42 · click to enlarge
Page 42

(5.3) Others

(1) Mixing the session of the TRT inquiry and the session of the test results.

A. General sequence

Rx Dx Dx Rx

AU2700/5400

Sx

Section Definition of section Limitation (a) The start of the transmission of the text of T.R.I. inquiry -- The end of the receiving of all T.R.I. for one samples

Don't transmit the Test result text.

(b) The start of the transmission of the text of the test results -- The end of the receiving of all result for one samples

Don't transmit the T.R.I. text.

Show 4 more labels / short text on this page
・・・ · Host · (a) (b) · (2) Detail
Page 43 of the 2Wire AU2700 user manual — 6. Specification of Connection
Page 43 · click to enlarge
Page 43

6. Specification of Connection

(6.1) In/Output signals and Terminals

PC Terminal Terminal at AU2700/5400 at host Signal Name Abbrv. Direction computer

(25 Pin Configuration) Ground FG - 1 Data sent TxDATA 3 3 (Rx DATA) Data receive RxDATA 2 2 (Tx DATA) Signal Ground SG 5 7 (SG)

Request to send RTS 7 Clear to send CTS 8

(6.2) Signal Level

Signal/Signal Format Signal Level

SPACE (ON) +3V and higher ("H") MARK (OFF) -3V and lower ("L")

(6.3) Applicable Connector

Style of connector : Confirm the standard of IBM PC (RS232 DB9 Male) Provide a RS232 DB9 Female connector. Style of connector Cover : ditto

(6.4) Others

A. When powering ON or OFF the host computer system, ensure that the AU2700/5400 is in Standy mode. Powering ON or OFF the host computer system while the AU2700/5400 is being operated may result in abnormality function of the AU2700/5400.

B. The cable for host communication is not included in the AU2700/5400 standard accessories. A shielded cable must be used for host connection.

C. Do not use the terminals on the AU2700/5400 other than No. 2,3,5,7, or 8 for any purpose.

Page 44 of the 2Wire AU2700 user manual — Y
Page 44 · click to enlarge
Page 44

A. Appendix

(1) Flags list Flag Contents ( Failed to evade the contamination since the detergent for evading the contamination is not enough. % The sample probe is clogged with sample. ? Calculation unable due to abnormal photometric data

# Sample level detection error R Reagent level detection error ! Calculation unable due to abnormal data

Reagent absorbance value at P-START of Reagent Blank run, is smaller than the Lower limit of the Parameter.

U

Reagent absorbance value at P-START is lower than the lower limit specified In the Parameters in routine run.

Reagent absorbance value at P-END of Reagent Blank run, is greater than the Upper limit of the Parameter.

Y

Reagent absorbance value at P-END is higher than the upper limit specified In the Parameters in routine run. @ Abnormally high result; absorbance of every wavelength is more than 2.5. $ No linearity validation conducted because less than 3 data obtained in the Kinetics. D Too quick reaction slope in increasing kinetics, absorbance at P-START is Higher than MAX. OD in increasing FIXED assay, or too slow reaction slope In decreasing kinetics (= no reaction observed) B Too quick reaction slope in increasing kinetics, or absorbance at P-END is Lower than MIN. OD in increasing FIXED assay. Linearity error in kinetics. * Linearity error in kinetics. & Prozone data error (prozone check point data error) Z Prozone error F Result higher than the dynamic range specified in the Parameters G Result lower than the dynamic range specified in the Parameters x Other QC data error 1 QC result beyond the QC value range specified in the Parameters 2 Multi-rule QC data error 3 Multi-rule QC data error 4 Multi-rule QC data error 5 Multi-rule QC data error 6 Multi-rule QC data error 7 Multi-rule QC data error p Result beyond the panic value specified in the Parameters T Abnormality found in the Inter-Item Check P Result higher than DECIDE RANGE designated in parameters. N Result lower than DECIDE RANGE designated in parameters. H Higher than the result value range specified in the Parameters L Lower than the result value range specified in the Parameters J Result higher than the repeat run range specified in the Parameters K Result lower than the repeat run range specified in the Parameters S Result extracted for repeat run

u · y

Page 45 of the 2Wire AU2700 user manual — 6. Specification of Connection
Page 45 · click to enlarge
Page 45

/ Test not performed: test has been requisitioned but not performed due To any reason. r transferred data to host e edited data c correct data by manual correction a On-board stability over b Calibration stability over or Calibration does not exist. ) Current reagent lot number was different from reagent lot No. when ACAL analysis

Page 46 of the 2Wire AU2700 user manual — 6. Specification of Connection
Page 46 · click to enlarge
Page 46

(2) Online parameters list Transmission information Contents menu T.R.I. receive

2Wire AU2700 — 6. Specification of Connection diagram (page 46)
Fig. 37 · 6. Specification of Connection

Normal Repeat sample Emergency Repeat sample Stat Repeat sample

Normal sample Emergency sample Stat sample Stat Fast sample

Real-time/batch/none Real-time/batch/none Real-time/batch/none Real-time/none Normal Repeat sample Emergency Repeat sample Stat Repeat sample

Real-time/batch/none Real-time/batch/none Real-time/batch/none QC sample Calibration sample Reagent blank sample

Real-time/batch/none Real-time/batch/none Real-time/batch/none Protocol 2 Contents menu Error control

[PARAMETER] [ONLINE] [PROTOCOL] menu. Protocol –1 Contents menu Data format

start code 1 start code 2 end code 1 end code 2 Max text length unit no. use ETB

4800/9600 class A/class B use/none (0-3) nn Text Format

bit/sec protocol BCC check retry times time out T1 T2 T3 T4 T5 T6 T7

Page 47 of the 2Wire AU2700 user manual — 6. Specification of Connection
Page 47 · click to enlarge
Page 47

Text information Contents menu rack no. /cup position rack no. digit

[PARAMETER] [ONLINE] [PROTOCOL] menu.

[PARAMETER] [FORMAT] [REQUISITION FORMAT] menu. patient information 1 patient information 2 patient information 3 patient information 4 patient information 5 patient information 6

use/none digits use/none digits use/none digits use/none digits use/none digits use/none digits

[PARAMETER] [FORMAT] [REQUISITION FORMAT] menu.

sample ID digits 4-26 digits [PARAMETER] [FORMAT] [REQUISITION FORMAT] menu. data format 6 /9 digits [PARAMETER] [ONLINE] [PROTOCOL] menu. data zero suppress used/none [PARAMETER] [ONLINE] [PROTOCOL] menu. Sample Type used/none [PARAMETER] [ONLINE] [PROTOCOL] [SPECIAL] menu. Dummy (Test Result Transmission Text)

used/none [PARAMETER] [ONLINE] [PROTOCOL] [SPECIAL] menu. Block Identification No. used/none [PARAMETER] [ONLINE] [PROTOCOL] [SPECIAL] menu. ANL Unit No. used/none Note 3 [PARAMETER] [ONLINE] [PROTOCOL] [SPECIAL] menu. Cuvette Side used/none Note 3 [PARAMETER] [ONLINE] [PROTOCOL] [SPECIAL] menu. Sequence of S.No. and S.ID in T.R.I text

S.No. / S.ID [PARAMETER] [ONLINE] [PROTOCOL] [SPECIAL] menu. Sequence of S.No. and S.ID in Test result text

S.No. / S.ID [PARAMETER] [ONLINE] [PROTOCOL] [SPECIAL] menu. Lot No. and Bottle No. used/none [PARAMETER] [ONLINE] [PROTOCOL] [SPECIAL] menu. Note 1)When you send the “Block Identification No.” field as none, the program will change the parameters automatically as following. End Code = ETX ETB Control = used Note 3) Only a Normal sample and repeat sample can be changed.

use/none 4/5 · Sex age/month · use/none use/none

Page 48 of the 2Wire AU2700 user manual — 6. Specification of Connection
Page 48 · click to enlarge
Page 48

Each of the following parameters are influenced as shown below.

2Wire AU2700 — 6. Specification of Connection diagram (page 48)
Fig. 38 · 6. Specification of Connection

RB R_ RH RE S_ SH SE DB D_ d_ DH dH DR DA DQ DE Sample Type - ○ × - ○ × - - ○ ○ × - Dummy (Test Result Transmission Text)

- - × - ○ × - - ○ ○ × -

- ○ × - × × - - ○ ○ × -

- - × - - × - × -

Sequence of S.No. and S.ID in Test result text

- ○ ○ × -

ANL Unit No. - ○ × × × - Cuvette Side - ○ × × × - Lot No. and Bottle No.

- ○ -

○influenced × : not influenced - : not apply
Page 49 of the 2Wire AU2700 user manual — (3) Character Table
Page 49 · click to enlarge
Page 49
2Wire AU2700 — (3) Character Table diagram (page 49)
Fig. 39 · (3) Character Table

HIGH LOW 0 1 2 3 4 5 6 7 8 9 A B C D E F 0 DE SP 0 @ P p 1 SH D1 ! 1 A Q a q 2 SX D2 " 2 B R b r 3 EX D3 # 3 C S c s 4 ET D4 $ 4 D T d t 5 EQ NK % 5 E U e u 6 AK SN & 6 F V f v 7 BL EB ' 7 G W g w 8 BS CN ( 8 H X h x 9 HT EM ) 9 I Y i y A LF SB * : J Z j z B H M

C CL FS , < L \ l | D CR GS - = M ] m } E SO RS . > N ^ n ~ F SI US / ? O _ o DL

From 20H to 7EH: These characters can be used for the Sample ID code. From 30H to 39H: These characters can be used for Test numbers. A1 - DC is Japanese character set.

(3) Character Table

Page 50 of the 2Wire AU2700 user manual — T2 [ ] (0-99)
Page 50 · click to enlarge
Page 50

(4) AU2700/5400 Online Parameter Sheet

Set up Setting T.R.I. Receive

Routine Normal † real time † batch † none Routine Emergency † real time † batch † none

Routine Repeat † real time † batch † none

Emergency Repeat † real time † batch † none Stat Normal † real time † batch † none Stat Repeat † real time † batch † none

Routine Normal † real time † batch † none Routine Emergency † real time † batch † none

Routine Repeat † real time † batch † none

Emergency Repeat † real time † batch † none

Stat Normal † real time † batch † none

Stat Repeat † real time † batch † none Stat Fast † real time † none

QC † real time † batch † none Calibration † real time † batch † none

Reagent blank † real time † batch † none

Upper protocol Setting Error control

T.R.I. Receive †continue † stop Results Transfer †continue † stop

Lower protocol Setting Character format

Character length. † 7 † 8 Parity bit † odd † even † none

Stop bit † 1 † 2

Start code 1 [ ] (01-1F) Start code 2 [ ] (00-1F)

End code 1 [ ] (01-1F)

End code 2 [ ] (00-1F)

Text length † 256 † 512 † 1024

Unit No. [ ] (00-99 space)

ETB control † yes † no

SEQ. Control

Bit/sec † 4800 † 9600 Class † Class A † Class B

BCC check † yes † no

Retry [ ] (0-3)

Time out T1 [ ] (0-99)

T2 [ ] (0-99)

T3 [ ] (0-99)

T4 [ ] (0-99)

T5 [ ] (0-99)

T6 [ ] (0-99)

T7 [ ] (0-99)

Results Transfer · Data format · Text format Setting

Page 51 of the 2Wire AU2700 user manual — (4) AU2700/5400 Online Parameter Sheet?
Page 51 · click to enlarge
Page 51

Rack No. † none † 4 † 5 Sex † yes † no AGE/MONTH † yes † no

patient inf. 1 † yes[ ]digits † no

patient inf. 2 † yes[ ]digits † no

patient inf. 3 † yes[ ]digits † no

patient inf. 4 † yes[ ]digits † no

patient inf. 5 † yes[ ]digits † no

patient inf. 6 † yes[ ]digits † no

ID digits [ ]digits (4-26) Data format † 6 † 9 ZERO Suppress † yes † no Sample Type † yes † no Dummy † yes † no Block Identification † yes † no ANL unit No. † yes † no Cuvette side † yes † no Sequence of S.No. and S.ID in T.R.I text

† No. † S.ID

Sequence of S.No. and S.ID in Test result text

† yes † no

Lot No. and Bottle No. † yes † no

Page 52 of the 2Wire AU2700 user manual — (5) ONLINE ALARM LIST
Page 52 · click to enlarge
Page 52

(5) ONLINE ALARM LIST

1)ONLINE ERROR [ aa ] [ bbbbbbbbb ] (1). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive has been designated as "STOP", then the following will occur: 1) Text following the alarm will not be received. 2) In a case when the text requisition is received real-time,the "STOP" status is ignored at the start of the next analysis and the text requisition will be received real-time again. (2). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive has been designated to "CONTINUE", the test requisition for the next sample will be received real-time. (1).Communication error occurs when transmittng the text for test requisition or when receiving the test requisition format. (2).The display in the [ ] and its meaning are as follows. aa : type of the errors 01 : Device name error 02 : Framing error 03 : Over run error 04 : Parity error 05 : Time out error 06 : Receive NAK when transmitting a text 07 : BCC error when receiving a text 08 : Other communication errors 09 : Function error (Full Duplex instead of Half Duplex) 10 : Unit name error 11 : Parameter error 12 : Request cancel bbbbbbbbb : Text having error RB : Start text for receiving the test requisition information. RE : End text for receiving the test requisition information. R_ __ nnnn : normal run, normal sample, sample No. R_ _E nnnn : normal run, emergency sample, sample No. R_ _P nnnn : normal run, STAT sample, sample No. R_ U_ nnnn : normal run, urine sample, sample No. R_ UE nnnn : normal run, urine emergency sample, sample No. R_ UP nnnn : normal run, urine STAT sample, sample No. R_ X_ nnnn : normal run, urine sample, sample No. R_ XE nnnn : normal run, urine emergency sample, sample No. R_ XP nnnn : normal run, urine STAT sample, sample No. RH __ nnnn : repeat run, normal sample, sample No. RH _E nnnn : repeat run, emergency sample, sample No. RH _P nnnn : repeat run, STAT sample, sample No. RH U_ nnnn : repeat run, urine sample, sample No. RH UE nnnn : repeat run, urine emergency sample, sample No. RH UP nnnn : repeat run, urine STAT sample, sample No. R_ __ xxxxxxxxxxx: normal run, normal sample, sample ID R_ _E xxxxxxxxxxx: normal run, emergency sample, sample ID R_ _P xxxxxxxxxxx: normal run, STAT sample, sample ID R_ U_ xxxxxxxxxxx: normal run, urine sample, sample ID R_ UE xxxxxxxxxxx: normal run, urine emergency sample,sample ID R_ UP xxxxxxxxxxx: normal run, urine STAT sample, sample ID R_ X_ xxxxxxxxxxx: normal run, urine sample, sample ID R_ XE xxxxxxxxxxx: normal run, urine emergency sample,sample ID R_ XP xxxxxxxxxxx: normal run, urine STAT sample, sample ID RH __ xxxxxxxxxxx: repeat run, emergency sample, sample ID RH _E xxxxxxxxxxx: repeat run, emergency sample, sample ID RH _P xxxxxxxxxxx: repeat run, STAT sample, sample ID

Page 53 of the 2Wire AU2700 user manual — (5) ONLINE ALARM LIST
Page 53 · click to enlarge
Page 53

RH U_ xxxxxxxxxxx: repeat run, urine sample, sample ID RH UE xxxxxxxxxxx: repeat run, urine emergency sample, sample ID RH UP xxxxxxxxxxx: repeat run, urine STAT sample, sample ID RH X_ xxxxxxxxxxx: repeat run, emergency sample, sample ID RH XE xxxxxxxxxxx: repeat run, emergency sample, sample ID RH XP xxxxxxxxxxx: repeat run, STAT sample, sample ID

Page 54 of the 2Wire AU2700 user manual — (5) ONLINE ALARM LIST
Page 54 · click to enlarge
Page 54

2)ONLINE ERROR [ aa ] [ bbbbbbbbb ] (1). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive is designated as "STOP", the following events can occur: 1). The text following the alarm will not be received. 2). In a case when the text requisition is received real-time,the "STOP" designation is ignored and the text requisition will be received in real-time at the start of the next analysis. (2). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive is designated as "Continue", the test requisition for the next sample will be received. (1). Communication error occurs when transmitting the text for test requisition or receiving the test requisition format. (2). The display in the [ ] and its meaning are as follows. aa : type of the errors 01 : Device name error 02 : Framing error 03 : Over run error 04 : Parity error 05 : Time out error 06 : Receive NAK when transmitting a text 07 : BCC error when receiving a text 08 : Other communication errors 09 : Function error (Full Duplex instead of Half Duplex) 10 : Unit name error 11 : Parameter error 12 : Request cancel bbbbbbbbb : Text having error RB : Start text for receiving the test requisition information. RE : End text for receiving the test requisition information. R_ __ nnnn : normal run, normal sample, sample No. R_ _E nnnn : normal run, emergency sample, sample No. R_ _P nnnn : normal run, STAT sample, sample No. R_ U_ nnnn : normal run, urine sample, sample No. R_ UE nnnn : normal run, urine emergency sample, sample No. R_ UP nnnn : normal run, urine STAT sample, sample No. R_ X_ nnnn : normal run, Other normal sample, sample No. R_ XE nnnn : normal run, Other emergency sample, sample No. R_ XP nnnn : normal run, Other STAT sample, sample No. RH __ nnnn : repeat run, normal sample, sample No. RH _E nnnn : repeat run, emergency sample, sample No. RH _P nnnn : repeat run, STAT sample, sample No. RH U_ nnnn : repeat run, urine sample, sample No. RH UE nnnn : repeat run, urine emergency sample, sample No. RH UP nnnn : repeat run, urine STAT sample, sample No. RH X_ nnnn : repeat run, Other normal sample, sample No. RH XE nnnn : repeat run, Other emergency sample, sample No. RH XP nnnn : repeat run, Other STAT sample, sample No. R_ __ xxxxxxxxxxx: normal run, normal sample, sample ID R_ _E xxxxxxxxxxx: normal run, emergency sample, sample ID R_ _P xxxxxxxxxxx: normal run, STAT sample, sample ID R_ U_ xxxxxxxxxxx: normal run, urine sample, sample ID R_ UE xxxxxxxxxxx: normal run, urine emergency sample, sample ID R_ UP xxxxxxxxxxx: normal run, urine STAT sample, sample ID R_ X_ xxxxxxxxxxx: normal run, Other normal sample, sample ID R_ XE xxxxxxxxxxx: normal run, Other emergency sample, sample ID R_ XP xxxxxxxxxxx: normal run, Other STAT sample, sample ID RH __ xxxxxxxxxxx: repeat run, normal sample, sample ID RH _E xxxxxxxxxxx: repeat run, emergency sample, sample ID

Page 55 of the 2Wire AU2700 user manual — (5) ONLINE ALARM LIST
Page 55 · click to enlarge
Page 55

RH _P xxxxxxxxxxx: repeat run, STAT sample, sample ID RH U_ xxxxxxxxxxx: repeat run, urine sample, sample ID RH UE xxxxxxxxxxx: repeat run, urine emergency sample, sample ID RH UP xxxxxxxxxxx: repeat run, urine STAT sample, sample ID RH X_ xxxxxxxxxxx: repeat run, Other normal sample, sample ID RH XE xxxxxxxxxxx: repeat run, Other emergency sample, sample ID RH XP xxxxxxxxxxx: repeat run, Other STAT sample, sample ID

Page 56 of the 2Wire AU2700 user manual — (5) ONLINE ALARM LIST
Page 56 · click to enlarge
Page 56

3)ONLINE FORMAT ERROR (1).Disregard the test requisition information text for the sample generating the alarm. (2).Continue to receive the test requisition information text for samples following the alarm. (1).In the Online Parameter menu, submenu Set Up, the T.R.I. receive is not designated asn "YES" for the necessary parameter. (2).The display in the [ ] and its meaning are as follows. aaaa : Type of error R_ __ : normal run, a normal sample R_ _E : normal run, an emergency sample R_ _P : normal run, a STAT sample R_ U_ : normal run, a urine normal sample R_ UE : normal run, a urine emergency sample R_ UP : normal run, a urine STAT sample R_ X_ : normal run, a Other normal sample R_ XE : normal run, an Other emergency sample R_ XP : normal run, a Other STAT sample RH __ : repeat run, a normal sample RH _E : repeat run, an emergency sample RH _P : repeat run, a STAT sample RH U_ : repeat run, a urine normal sample RH UE : repeat run, a urine emergency sample RH UP : repeat run, a urine STAT sample RH X_ : repeat run, a Other normal sample RH XE : repeat run, an Other emergency sample RH XP : repeat run, a Other STAT sample bbbbbbbbbb : Information type that is not designated No sample No. No ID No.

Page 57 of the 2Wire AU2700 user manual — (5) ONLINE ALARM LIST
Page 57 · click to enlarge
Page 57

4)ONLINE ILLEGAL TEXT CODE [ aa ] (1).Disregard the test requisition information text for the sample generating the alarm. (2).Continue to receive the test requisition information text for samples following the alarm. (1).When receiving the test requisition information text using online, the text classification code is out of specification. (2).The display of [ ] and its meaning are as follows. aa : R_ : Normal sample RH : Repeat run sample 5)ONLINE ILLEGAL TEXT BLOCK No. [ aa -> bb ] (1). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive is designated as "STOP", the following events occur: 1).Disregard the test requisition information text for the sample generating the alarm. 2).System will not receive text for samples following the sample generating the alarm. 3).In a case when the text requisition is received in real-time, the "STOP" is ignored and the text requisition text will be received with real-time again during the next analysis. (2). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive is designated as "Continue", the following events occur: 1).Disregard the test requisition information text for the sample generating the alarm. 2).The test requisition information text for the next sample will be received. (1).When receiving the requisition information text using online, the block identification No. is out of specification. (2).The display of [ ] and its meaning are as follows. aa : Previously received block identification No. bb : Last received block identification No.

Page 58 of the 2Wire AU2700 user manual — (5) ONLINE ALARM LIST
Page 58 · click to enlarge
Page 58

6)ONLINE ILLEGAL SAMPLE No. [ aa: bbbbb ] (1).Disregard the test requisition information text for the sample generating the alarm. (2).Continue to receive the test requisition information text for other samples within the run. (1).When receiving the test requisition information text using online, sample No. of the sample generating the alarm does not meet the specification. (2).The display of [ ] and its meaning are as follows. aa : Sample type __ : Normal sample _E : Emergency sample _P : Stat sample U_ : Urine normal sample UE : Urine emergency sample UP : Urine stat sample X_ : Other Normal sample XE : Other Emergency sample XP : Other Stat sample bbbb : Sample No. 0001-9999 : Normal sample E001-E999 : Emergency P001-P999 : Stat sample

Page 59 of the 2Wire AU2700 user manual — (5) ONLINE ALARM LIST
Page 59 · click to enlarge
Page 59

7)ONLINE ILLEGAL RACK No. [ aaaaaaa: b: cccccccc ] (1). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive is designated as "STOP", the following events occur: 1).Disregard the test requisition information text for the sample generating the alarm. 2).System will not receive text for samples following the sample generating the alarm. 3).In a case when the text requisition is received in real-time, the "STOP" is ignored and the text requisition text will be received with real-time again during the next analysis. (2). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive is designated as "Continue", the following events occur: 1).Disregard the test requisition information text for the sample generating the alarm. 2).The test requisition information text for the next sample will be received. (1).When receiving the requisition information text using online, the rack No. is out of specification. (2).The display of [ ] and its meaning are as follows. aaa : Transferred sample type and sample No. __ nnnn : Normal sample _E nnnn : Emergency sample _P nnnn : Stat sample U_ nnnn : Urine normal sample UE nnnn : Urine emergency sample UP nnnn : Urine stat sample X_ nnnn : Other normal sample XE nnnn : Other emergency sample XP nnnn : Other Stat sample bbb : Received sample type and sample No. __ nnnn : Normal sample _E nnnn : Emergency sample _P nnnn : Stat sample U_ nnnn : Urine normal sample UE nnnn : Urine emergency sample UP nnnn : Urine stat sample X_ nnnn : Other normal sample XE nnnn : Other emergency sample XP nnnn : Other Stat sample b : Received the kind of rack N : Normal sample rack E : Emergency rack R : Repeat run rack ccccccc : Received rack No. and position in rack

Page 60 of the 2Wire AU2700 user manual — (5) ONLINE ALARM LIST
Page 60 · click to enlarge
Page 60

8)ONLINE ILLEGAL SAMPLE KIND No. [ aa ] (1).Disregard the test requisition information text for the sample generating the alarm. (2).Continue to receive the test requisition information text for the other samples within the run. (1).When receiving the test requisition information text using online, sample kind No. of the sample generating the alarm does not meet the specification. (2).The display of [ ] and its meaning are as follows. aa : Sample type __ : Normal sample _E : Emergency sample _P : Stat sample U_ : Urine normal sample UE : Urine emergency sample UP : Urine stat sample X_ : Other normal sample XE : Other emergency sample XP : Other stat sample 9)ONLINE ILLEGAL GENDER TEXT [ a ] (1).Disregard the test requisition information text for the sample generating the alarm. (2).Continue to receive the test requisition information text for other samples within the run. (1).When receiving the requisition information text using online, the sex of this text is not within the specification. (2).The display of [ ] and its meaning are as follows. a : Received sex 10)ONLINE ILLEGAL AGE/MONTH [ aaa bb ] (1).Disregard the test requisition information text for the sample generating the alarm. (2).Continue to receive the test requisition information text for other samples within the run. (1).When receiving the requisition information text using online, age or month age (age for baby before one year old) of this text is out of the specification. (2).The display of [ ] and its meaning are as follows. aaa : Received age bb : Received month age

Page 61 of the 2Wire AU2700 user manual — (5) ONLINE ALARM LIST
Page 61 · click to enlarge
Page 61

11)ONLINE ANALYSIS METHOD MISMATCH [ aa bb ] (1).Disregard the test requisition information text for the sample generating the alarm. (2).Continue to receive the test requisition information text for other samples within the run. (1).When receiving the test requisition information text using online, sample type of this text does not match the required sample type. (2).The display of [ ] and its meaning are as follows. aa : Sample type __ : Normal sample _E : Emergency sample _P : Stat sample U_ : Urine normal sample UE : Urine emergency sample UP : Urine stat sample X_ : Other normal sample XE : Other emergency sample XP : Other Stat sample bb : Received sample type __ : Normal sample _E : Emergency sample _P : Stat sample U_ : Urine normal sample UE : Urine emergency sample UP : Urine stat sample X_ : Other normal sample XE : Other emergency sample XP : Other Stat sample

12)ONLINE SAMPLE No. MISMATCH [ aaaaaa bbbbbb ]

Page 62 of the 2Wire AU2700 user manual — (5) ONLINE ALARM LIST
Page 62 · click to enlarge
Page 62

(1).Disregard the test requisition information text for the sample generating the alarm. (2).Continue to receive the test requisition information text for the other samples within the run. (1).When receiving the test requisition information text using online, sample type of this does not match the required sample. (2).The display of [ ] and its meaning are as follows. aaaaaaa : Transferred sample type and sample No. __ nnnn : Normal sample _E nnnn : Emergency sample _P nnnn : Stat sample U_ nnnn : Urine normal sample UE nnnn : Urine emergency sample UP nnnn : Urine stat sample X_ nnnn : Other normal sample XE nnnn : Other emergency sample XP nnnn : Other Stat sample bbbbbb : Received sample type and sample No. __ nnnn : Normal sample _E nnnn : Emergency sample _P nnnn : Stat sample U_ nnnn : Urine normal sample UE nnnn : Urine emergency sample UP nnnn : Urine stat sample X_ nnnn : Other normal sample XE nnnn : Other emergency sample XP nnnn : Other Stat sample

Page 63 of the 2Wire AU2700 user manual — (5) ONLINE ALARM LIST
Page 63 · click to enlarge
Page 63

13)ONLINE RACK No. MISMATCH [ aaaaaa : bbbb cccc ] (1). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive is designated as "STOP", the following events occur: 1).Disregard the test requisition information text for the sample generating the alarm. 2).System will not receive text for samples following the sample generating the alarm. 3).In a case when the text requisition is received in real-time, the "STOP" is ignored and the text requisition text will be received with real-time again during the next analysis. (2). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive is designated as "Continue", the following events occur: 1).Disregard the test requisition information text for the sample generating the alarm. 2).The test requisition information text for the next sample will be received. (1).When receiving the test requisition information text using online, rack No. of this does not match the required sample. (2).The display of [ ] and its meaning are as follows. aaaaaaaaa : Transferred sample type and sample No. __ nnnn : Normal sample _E nnnn : Emergency sample _P nnnn : Stat sample U_ nnnn : Urine normal sample UE nnnn : Urine emergency sample UP nnnn : Urine stat sample X_ nnnn : Other normal sample XE nnnn : Other emergency sample XP nnnn : Other Stat sample bbbb : Transferred rack No. cccc : Received rack No. 14)ONLINE MISMATCH [ aaaaaaaaaa bbbbbbbbbbb ] (1).Disregard the test requisition information text for the sample generating the alarm. (2).Continue to receive the test requisition information text for the other samples within the run. (1).When receiving the test requisition information text using online, sample ID of this text does not match the required ones. (2).The display of [ ] and its meaning as follows. aaaaaaaaa : Transferred sample ID bbbbbbbbb : Received sample ID

Page 64 of the 2Wire AU2700 user manual — (5) ONLINE ALARM LIST
Page 64 · click to enlarge
Page 64

15)ONLINE ITEM ERROR [ aaaaaaaa ] (1).Disregard the test requisition information text for the sample generating the alarm. (2).Continue to receive the test requisition information text for the other samples within the run. (1).When receiving the test requisition information text using online, the test requisition cannot be stored. (2).The display of [ ] and its meaning are as follows. aaaaaa : Sample type and sample No. __nnnn : Normal sample __Ennn : Emergency sample __Pnnn : Stat sample _U_nnn : Urine normal sample _UEnnn : Urine emergency sample _UPnnn : Urine stat sample _Xnnnn : Other normal sample _XEnnn : Other emergency sample _XPnnn : Other Stat sample H_nnnn : repeat run normal sample H_Ennn : repeat run emergency sample H_Pnnn : repeat run stat sample HUnnnn : repeat run urine normal sample HUEnnn : repeat run urine emergency sample HUPnnn : repeat run urine stat sample HXnnnn : repeat run other normal sample HXEnnn : repeat run other emergency sample HXPnnn : repeat run other stat sample

Page 65 of the 2Wire AU2700 user manual — (5) ONLINE ALARM LIST
Page 65 · click to enlarge
Page 65

16)ONLINE RERUN ITEM ERROR [ aaaaaaaa ] (1).Disregard the repeat run test requisition information for the sample generating the alarm. (2).Continue to receive the test requisition information text for the other samples within the run. (1).When receiving the test requisition information text for repeat run using online, repeat run test requisition cannot be stored with for the following reasons. 1).Original sample that is designated in the repeat run text is not stored in the normal run/sample test requisition information. 2).Sample No. is not set to the original sample that is designated in the repeat run text. 3).Original sample that is designated in the repeat run text has already been stored as a repeat run sample for another sample. (2).The display of [ ] and its meaning are as follows. aaaaaa : sample type and original sample No. __nnnn : Normal sample __Ennn : Emergency sample __Pnnn : Stat sample _Unnnn : Urine normal sample _UEnnn : Urine emergency sample _UPnnn : Urine stat sample _Xnnnn : Other normal sample _XEnnn : Other emergency sample _XPnnn : Other Stat sample

Ask about the 2Wire AU2700

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the 2Wire AU2700

Be the first to ask — other owners or our team will answer.

Original document

2Wire AU2700 manual PDF

65 pages·389 KB·Coffeemaker·PDF
Click any page above to open it here
Page 1 / …
100%
Loading pages…