Software Release Notes & Installation Instructions Tektronix, Inc. 14150 SW Karl Braun Drive P.O. Box 500 Beaverton, OR 97077 USA
These release notes contain important information about version 1.6.1 of the AFG31000 software.
This document provides supplemental information regarding the behavior of the AFG31000 software. This information is grouped into six categories:
Revision history Lists the version of software, the document version, and the date of the software release.
New features / enhancements Summary of each significant new feature included.
Problem fixes Summary of each significant software/firmware bug fix
Known problems Description of each significant known problem and ways to work around it.
Installation instructions Detailed instructions describing how to install the software.
Appendix A – Previous versions Contains information about previous versions of the software.
This document is periodically updated and distributed with releases and service packs to provide the most up-to-date information. This revision history is included below.
3/23/2021 V1.6.1 0771639 02
12/3/2020 V1.6.0 0771639 01
9/30/2019 V1.5.2 0771639 00
11/15/2018 V1.4.6 - -

Problem fixes

New features/enhancements



Symptom The predefined equations in ArbBuilder will not compile with the default settings.
Workaround Change the range or the number of points for the equation to compile properly.
Symptom When running the Refresh Relay action using the Utility front panel hard key, the display operations are not locked out, allowing other functions to run.
Workaround It is recommended that the Refresh Relay action be run using the touch screen menus. If it is run using the Utility front panel hard key, then do not select any other options from the touch screen until the action completes.

You can use the front-panel USB Type A connector to update your instrument firmware using a USB flash drive. This task is done using the front-panel touchscreen.
CAUTION. Updating your instrument firmware is a sensitive operation; it is critical that you follow the instructions below. If you do not, you may cause damage to your instrument. For example, to prevent damage to the instrument, do not remove the USB flash drive at any time while updating the firmware, and do not power off the instrument during the update process. To update your instrument firmware: 1. Visit tek.com and search for the Series 31000 firmware.








Symptom The predefined equations in ArbBuilder will not compile with the default settings.
Workaround Change the range or the number of points for the equation to compile properly.
Models affected AFG31151, AFG31152, AFG31251, and AFG31252
Enhancement Support the AFG31151, AFG31152, AFG31251, and AFG31252 models.
Models affected AFG31151, AFG31152, AFG31251, and AFG31252
Enhancement Optimized user interface.
Issue number 1 · Issue number 2
Common problems, likely causes, and what to do.
| Problem | Possible cause | Remedy | |
|---|---|---|---|
| Models affected | p.8 | ||
| Symptom | p.8 | ||
| Resolution | p.8 | ||
| Issue number | AFG-308 / AR62443 AFG31XXX This update fixes an issue setting the generated waveform using the Recall feature in Basic mode. The pulse width was not always set properly, causing unexpected results. This issue has been corrected. | p.8 | |
| Issue number | AFG-310 / AR62352 AFG31XXX The user will not get the expected waveform when they attempt an AM modulation with an Arb file greater than 4,096 points. The max points for AM modulation using an Arb waveform is 4,096 points. This issue has been corrected. The product datasheet has been updated. | p.8 | |
| Issue number | AFG-316 / AR62581 AFG31XXX Undesired glitches could occur in burst mode idle state or when turning the output on and off. This issue has been corrected. | p.8 | |
| Issue number | AFG-324 AFG31XXX Instrument Ethernet connection using DHCP mode is unstable, repeatedly disconnecting and reconnecting, for a long time with certain network configurations. This issue has been corrected. | p.8 | |
| Issue number | AFG-330 AFG31XXX Grammar and typographical errors in the automatic calibration dialog. This issue has been corrected | p.8 | |
| Issue number | AFG-337 AFG31XXX Grammar and typographical errors in the self-diagnostics dialog. This issue has been corrected. | p.8 | |
| Issue number | AFG-352 / AR62937 AFG31XXX In Sequence mode, the idle value of the signal is always the offset of the waveform (or example, 2.5 V of a 0 to 5 Vpp waveform), This will ultimately distort the customer's desired waveform shape. Changed the default sequence mode from idle value to be 0 V if the waveform can reach 0 V. Otherwise the idle value will be the offset. | p.8 |