Handheld Density Meter
SG-Pro











4 Table of Contents Eagle Eye Power Solutions – SG-Pro

Thank you for choosing an Eagle Eye Power Solutions density meter. The handheld density meters SG-Pro and DensitoPro are handheld measuring instruments used to determine the density of liquids.
The instructions in this document refer to handheld density meters SG-Pro and DensitoPro running firmware version V1.0.0 or higher.
The firmware license is subject to the End User License Agreement EULA version 3.0.
If you have any additional questions, contact your authorized EEPS dealer or service represen- tative.
Refers to an external document.
Elements of instructions ▪ Prerequisites 1 Steps 2 ...
See also View the firmware version and other system information } Page 61
Introduction 5 Eagle Eye Power Solutions - SG-Pro
About this document · Conventions and symbols · Intermediate results Results

Only use the instrument according to the Reference Manual. If you do not use the instrument according to the Reference Manual or if it is modified, the safety of the instrument may be impaired and EEPS assumes no liability.
Safety notes contain important information on safety issues. Ignoring the safety notes may lead to personal injury, damage to the instrument, malfunctions and false results. Safety notes are marked with the following signal words and warning symbols:
The density meter is designed to be used by trained staff and is intended for measuring the density of liquid samples that are compatible with the materials with which they come into contact.
The density meter is designed for indoor and outdoor operation in dry conditions. The following site requirements apply:
Any other type of use and operation beyond the limits of use stated by EEPS without consent from EEPS is considered as not intended.
This device complies with Industry Canada licence-exempt RSS standard(s) and part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause interference, and (2) this device must accept any interference, including interference that may cause undesired operation of the device. Changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment.
Responsibilities of the instrument owner
The instrument owner is the person holding the legal title to the instrument and who uses the instrument or authorizes any person to use it, or the person who is deemed by law to be the operator of the instrument. The instrument owner is responsible for the safety of all users of the instrument and third parties.
EEPS assumes that the instrument owner trains users to safely use the instrument in their workplace and deal with potential hazards. EEPS assumes that the instrument owner provides the necessary protective gear.
6 Safety Information Eagle Eye Power Solutions - SG-Pro
Signal words · Electrical shock · Intended use

Danger of death or serious injury due to electric shock! Contact with parts that carry a live current can lead to death or injury. 1 Only use the AC adapter designed for your instrument. 2 Keep all electrical cables and connections away from liquids and moisture. 3 Check the cables and the plugs for damage and replace damaged cables and plugs.
Risk of damage to the instrument due to the use of unsuitable parts! Using unsuitable parts with the instrument can damage the instrument or cause it to malfunction.
−Only use parts from EEPS that are intended to be used with your instrument.
Safety Information 7 Eagle Eye Power Solutions - SG-Pro
Safety notes

3.1 Instrument overview
1 2 3 11 4 10


No. Name Function
1 USB-C socket To connect the density meter with the AC adapter or a computer.
2 RFID reader and barcode reader (DensitoPro only )
To scan a linear barcode or read and write information to Smart Tags.
3 Battery cover To access the battery.
4 USB-A socket To connect a USB printer or bluetooth dongle.
5 Sample pump cover To access the sample pump.
6 Sample pump window
To view the sample pump and to check if the sample pump is filled or empty.
7 Filling tube To fill and drain the measuring cell.
8 Filling opening with locking screw
To fill and drain the measuring cell using a syringe. The locking screw closes off the filling opening.
9 Measuring cell window
To view the measuring cell and to check if the measuring cell is filled or empty.
10 Navigation joystick To navigate through the menus.
11 Key 1 To start up the density meter, select menu items and to navigate.
12 Key 2 To select menu items and to navigate.
13 Key 3 To select menu items and to navigate.
14 Screen To display settings and results.
The density meter uses the oscillating body method as measurement technology. The density meter has no active temperature control.
8 Design and Function Eagle Eye Power Solutions - SG-Pro
13 12 14

If you measure samples with low viscosity, you can use the sample pump (1) and the filling tube (2) to fill the measuring cell (5). For samples with high viscosity you can use a syringe (3) and the connector for the syringe (4) to fill the measuring cell (5). The user can observe the filling through the window of the measuring cell.



For repeated measurements, you can define and save workflows as methods. Two workflow types are available:
The results are automatically converted into one of the following predefined units or a unit that the user defines.
Design and Function 9 Eagle Eye Power Solutions - SG-Pro


d [g/cm3]
Temp. 23 °C
No. Name Explanation
1 Status ribbon Displays the current date, time, battery charge status and connection to acces- sories like EasyDirect Density & Refractometry.
2 Method ribbon Displays the type of density determination and the method identification or the selected standard.
3 Sample ID ribbon Displays the sample ID.
4 Measuring ribbon Displays the unit, the measured value or result and the temperature of the measuring cell.
5 Sampling ribbon Displays the current status of the sample pump or syringe.
6 Key functions ribbon Displays the function of the keys.
Status ribbon Barcode reader is scanning. Status ribbon RFID reader is reading or writing.
Status ribbon EasyDirect Density & Refractometry is connected.
Status ribbon Density meter is sending data to the printer.
Status ribbon Shows the charge of the battery.
Status ribbon Battery is charging.
Sampling ribbon Syringe mode activated. The sample pump is deactivated.
Sampling ribbon Sample pump activated and piston in the lowermost position.
Sampling ribbon Sample pump activated and piston in the uppermost position.
10 Design and Function Eagle Eye Power Solutions - SG-Pro
Menu Results Start · Icon Location Explanation

4.1 Scope of delivery
Filling Tube 190 mm
Power Supply and Worldwide Adapters
Tag SmartSample (10 pcs) 30449268 – •
1 Remove the density meter from the protective packaging. 2 Store the packing material for later transport over long distances. 3 Check if you received all parts listed in the scope of delivery. 4 Inspect the parts visually for flaws or damage. 5 If parts are missing or damaged, report it immediately and file a freight claim if needed.

You can either charge the density meter using the supplied AC adapter or a computer. Recharging an empty battery takes approximately three hours.
The AC adapter is suitable for all supply line voltages ranging from 100...240 V AC and 50/60 Hz.
Danger of death or serious injury due to electric shock! Contact with parts that carry a live current can lead to death or injury. 1 Only use the AC adapter designed for your instrument. 2 Keep all electrical cables and connections away from liquids and moisture. 3 Check the cables and the plugs for damage and replace damaged cables and plugs.
Assemble the AC adapter
1 Slide the desired set of prongs (1) into the AC adapter (2). 1

2 Insert the USB-A plug of the USB-C cable in the USB-A socket of the AC adapter.
Connect the density meter to the power supply
1 Insert the USB-C plug of the USB-C cable in the USB-C socket of the density meter. 2 Install the cables in such a way that they cannot be damaged or interfere with operation. 3 Insert the plug of the AC adapter in a power outlet that is easily accessible.
The density meter starts charging and the icon is displayed.
Change the prongs of the AC adapter
1 Pull the plug of the AC adapter out of the power outlet. 1 2 Push the button (2) and slide the set of prongs (1) out. 2
3 Slide the desired set of prongs into the AC adapter.

3 Insert the USB-A plug in the USB-A socket of the computer.
The density meter starts charging and the icon is displayed.
1 Press the Menu key.
2 Navigate to Settings > Measurement > Sampling.
The Sampling window opens.
3 Navigate to Sample pump.
4 Press the OK key. 5 Navigate back to the home screen. The icon is displayed in the sampling ribbon.


opening (3) and tighten it.
1 Press the Menu key.
2 Navigate to Settings > RFID / Barcode > Type.
The Type window opens.
3 To activate the RFID reader, navigate to RFID.
4 To deactivate the RFID reader, navigate to None.
See also Read and write information with the RFID reader (DensitoPro only) } Page 44
1 Press the Menu key.
2 Navigate to Settings > RFID / Barcode > Type.
The Type window opens.

3 To activate the barcode reader, navigate to Barcode.
4 To deactivate the barcode reader, navigate to None.
See also Read information with the barcode reader (DensitoPro only) } Page 45
When you work with a syringe you can place the density meter either on the bench or hold it upright.


2 Navigate to Settings > Measurement > Sampling.
The Sampling window opens.
3 Navigate to Syringe.
4 Press the OK key. 5 Navigate back to the home screen. The icon is displayed in the sampling ribbon. 6 If you want place the density meter on the bench, activate the auto rotation of the screen.
See also Activate and deactivate the auto rotation of the screen } Page 20


2 Screw the connector into the filling opening and tighten it.

See also Accessories } Page 65 Install the filling tube } Page 13
In order to print the results right after the measuring procedure, the density meter has to be connected to a USB or Bluetooth printer. Alternatively the results can be printed at a later time from the Results menu.
The following printer types are compatible with the density meter:
Bluetooth Godex MX20 USB P25
The bluetooth function is only available in selected countries with the necessary approvals. For more infor- mation, contact an authorized EEPS dealer or service representative.
Connection type Printer

Refer to the user documentation of your printer for setting up and operating the printer.
See also Accessories } Page 65
1 Press the Menu key.
2 Navigate to Settings > Printer > Printer type.
The Printer type window opens.
3 Navigate to USB.
5 Connect the USB-A plug of USB-C cable with the USB-A socket of the printer. 6 Connect the USB-A plug with the USB-A socket of the density meter. 7 Switch on the printer. The connection is established.
See also Configure the automatic printing of results } Page 31 Print results } Page 47
2 Press the Menu key.

3 Navigate to Settings > Printer > Printer type.
The Printer type window opens.
4 Navigate to Bluetooth.
6 Press the Menu key.
7 Navigate to Printer selection.
The Printer selection window with a list of
available bluetooth printers opens. 8 Navigate to the bluetooth printer you want to connect.
Configure the automatic printing of results } Page 31 Print results } Page 47
2 Navigate to Settings > Printer > Test printer.
If there is no connection yet, the notification Connecting to printer is displayed. When the connection is established, a test page is printed.
See also

See also Connect a printer } Page 15
EasyDirect Density & Refractometry is a data management software to manage a database of measurement results. By connecting the density meter via a USB-C cable to a computer on which EasyDirect Density & Refractometry is installed, the data can be directly exported to the software.
Operating EasyDirect Density & Refractometry is described in the help of the software.
See also Export results to EasyDirect Density & Refractometry } Page 47
The following date formats are available:
1 Press the Menu key.
2 Navigate to Settings > Date / Time.
The Date / Time window opens.
3 Navigate to Date or Time.
4 To change the date or time, move the navigation joystick up to raise or down to lower the number. 5 To switch between day, month and year or hours and minutes, move the navigation joystick to the left or right.
7 Navigate to Date format or Time format.
8 Navigate to the format you want to use.
Procedure · Procedure

Two temperature units are available.
When you change the temperature unit, temperatures are displayed in the selected unit. The temperature unit of results that were determined before you changed the unit remain unchanged.
1 Press the Menu key.
2 Navigate to Settings > Global units > Temperature unit.
The Temperature unit window opens.
3 Navigate to the desired Temperature unit.
The temperatures are displayed in the selected unit.
Three density units are available.
When you change the density unit, density values are displayed in the selected unit. There are two exceptions.
1 Press the Menu key.
2 Navigate to Settings > Global units > Density unit.
The Density unit window opens.
3 Navigate to the desired Density unit.
4 Press the OK key. The densities are displayed in the selected unit
The settings of the parameters Brightness, Energy saver and Auto shutdown influence the power consumption and thus the battery life.
Brightness Defines the brightness of the screen. 10...100 %
Energy saver Defines if the screen is dimmed when no action has been performed for a given time.
Off | 30 | 60 | 90 | 120 seconds
Auto shutdown Defines if the density meter is shut down when no action has
Off | 2 | 5 | 10 minutes
been performed for a given time.
1 Press the Menu key.
2 Navigate to Settings > Screen.
The Screen window opens.
3 Navigate to the desired parameter and change the settings.

With the parameter Password you can prevent users from performing certain actions without a password. The following table shows which actions are password-protected.
Start methods. No Start tests. No
Start adjustments. Yes View the configuration of methods, tests and adjustments. No
Change the configuration of methods, tests and adjustments. Yes
Create and delete methods. Yes
View, print and export results. No Delete all results. Yes
View the settings of the density meter. No Change the settings of the density meter. Yes
Update the firmware. Yes Restore factory adjustment. Yes
Restore factory settings. Yes
The password has to be entered once before a user performs a password-protected action and does not have to be reentered until the density meter is restarted. If the user forgets the password, the universal password 8606 unlocks all actions and settings.
1 Press the Menu key.
2 Navigate to Settings > Password > Password protection.
The Password protection window opens.
3 Navigate to Yes.
The Password window opens.
5 Enter a password with the navigation joystick.
6 Press the OK key. 7 Reenter the password.
The password has to be entered before the user can perform a protected action.
1 Press the Menu key.
2 Navigate to Settings > Password > Password protection.
The Password protection window opens.
3 Navigate to No.
The Password window opens. 5 Enter the password with the navigation joystick.
6 Press the OK key. Users can perform all actions.
Action Password protected

1 Disable the password protection. 2 Enable the password protection and enter the new password.
The Auto rotate parameter defines whether the screen rotates by 180° when the density meter is placed on a bench. 1 Press the Menu key.
2 Navigate to Settings > Screen > Auto rotate.
The Auto rotate window opens.
3 Navigate to Yes to switch on or No to switch off the Auto rotate mode.
4 Press the OK key. The screen rotates when the density meter is placed on a bench.
The parameter Audio signal defines whether an audio signal is played when a key or the navigation joystick is pressed and when measurements are finished. 1 Press the Menu key.
2 Navigate to Settings > Audio signal.
The Audio signal window opens.
3 Navigate to Yes to switch on or No to switch off the Audio signal.
1 Press the Menu key.
2 Navigate to Settings > Language / Keyboard > Keyboard layout.
The Keyboard layout window opens.
3 Navigate to the desired keyboard layout.
The system language defines the language of the user interface and printouts.
1 Press the Menu key.
2 Navigate to Settings > Language / Keyboard > Language.
The Language window opens. 3 Navigate to the desired language.
4 Press the OK key. The user interface is changed to the selected language.
Because the measuring cell always contains residue from the last sample, there is a risk that the measuring cell can be damaged by the sample. To prevent such damages, EEPS recommends that you clean the measuring cell at the end of each sample series.
To remind users to clean the measuring cell, you can activate a reminder. This reminder is displayed before manual shutdown of the instrument, and users have to confirm or cancel the reminder before shutting down the density meter.

Action of the system Cleaning reminder Auto shutdown
When users shut down the density meter, a reminder is displayed.
If the density meter automatically shuts down, no reminder is displayed.
No reminder is displayed when the density meter shuts down.
No Yes/No
1 Press the Menu key.
2 Navigate to Settings > Measurement > Cleaning reminder.
The Cleaning reminder window opens.
3 Navigate to Yes to activate or No to deactivate the Cleaning reminder.
See also Change energy saver settings } Page 18
Three different types of density determinations are available.
4.9.1.1 Create a method
The Methods window opens.
2 Press the New key.
The Method settings window opens.
3 Navigate to Method ID and enter a name with up to 25 characters.
The method is listed in the Methods window. 5 Change the parameters as needed.
See also Configure the user guidance } Page 23 Configure the measurement reliability } Page 24 Configure the result conversion } Page 24 Configure the automatic documentation of results } Page 30 Configure result limits } Page 32
Yes No · Yes Yes · Procedure

The Methods window opens.
2 Navigate to the method you want to edit.
The Method settings window opens. 3 Change the parameters as needed.
See also Configure the user guidance } Page 23 Configure the measurement reliability } Page 24 Configure the result conversion } Page 24 Configure the automatic documentation of results } Page 30 Configure result limits } Page 32
The Methods window opens. 2 Navigate to the method you want to delete.
The Method settings window opens.
3 Press the Delete key.
The message Do you want to delete the method? is displayed.
4 Navigate to Yes.
5 Press the OK key. The method is no longer listed in the Methods window.
1 Press the Menu key.
2 Navigate to Test.
The Test window opens.
3 Change the parameters as needed.
See also Configure the user guidance } Page 23 Configure the automatic documentation of results } Page 30 Configure a standard } Page 32 Configure the tolerance for a test } Page 33
1 Press the Menu key.
2 Navigate to Adjustment.
The Adjustment window opens.
3 Change the parameters as needed.

See also Configure the user guidance } Page 23 Configure the automatic documentation of results } Page 30 Configure a standard } Page 32
The parameters that are available to configure the user guidance are described in the following chapters.
The Workflow parameter defines the level of user guidance during the density determination. Two options are available.
The parameter Fill speed defines the speed of the pump for filling the measuring cell. The sample viscosity places limits on the fill speed.
Speed level % of maximum pump speed
Fill time for water Viscosity
High 80 % Approx. 4 s Up to 40 mPa*s
Medium 40 % Approx. 8 s Up to 75 mPa*s
Low 10 % Approx. 30 s Up to 100 mPa*s
The parameter Fill cycle defines, how many times the measuring cell is filled and emptied before the density is measured. The sample that is in the measuring cell after the last filling is used for the measurement. Recommendation:
Two types of cleaning methods are available, Drain and Rinse.
Drain The user is prompted to drain the measuring cell after the measurement. This setting is useful if the samples that are measured dissolve in each other.
The user is prompted to fill and drain the measuring cell with a cleaning solution after the measurement. The parameter Rinse cycle defines how many times the user is prompted to fiil and drain the measuring cell. This setting is only useful if the samples that are measured do not dissolve in each other.
See also Typical phases of density determinations } Page 36
Rinse

After the measurement reliability is achieved, the results are frozen. The parameter Confirmation of results defines whether users have to confirm the results before they are replaced by the current density value. Regardless of the setting, all results are stored. The setting is not method specific but valid for all density determinations.
Action of the system Confirmation of results Workflow
The screen is frozen and the sample pump is locked until users confirm the result.
The screen is frozen until users confirm the result but the sample pump is not locked.
Users are prompted to place the filling tube into the waste beaker. As soon as users press OK, the result is replaced by the current density value.
As soon as the user starts draining the cell, the result is replaced by the current density value.
1 Press the Menu key.
2 Navigate to Settings > Measurement > Confirmation of results.
The Confirmation of results window opens.
3 Navigate to Yes to activate or No to deactivate the Confirmation of results.
See also Define the level of user guidance } Page 23
Measurement reliability is only available for methods.
The parameter Measurement reliability defines how the stability of a measured value is evaluated. Once a measured value is determined to be stable, it is saved as a result.
Define the level of user guidance } Page 23
The density meter converts the measured oscillation to the unit you define. The conversions are grouped in the following categories.

Select the category and the conversion type
1 Press the Menu key.
2 Navigate to Methods > Method settings > Calculation > Category
3 Navigate to the desired category.
The Calculation window opens. 5 Navigate to the conversion type you want to use.
6 Press the OK key. The parameters that you need to configure for the selected conversion type are displayed. You can find more information about the conversion types and the parameters in the following chapters.
Available conversions:
Density at the measurement temperature
Unit: [g/cm3], [kg/m3] or [lb/gal (US)] as defined in Settings > Global units. The calculation is based on a reference for the density of water.
Unit: [g/cm3], [kg/m3] or [lb/gal (US)] as defined in Settings > Global units.
The measured density is converted to the value it would have at a compensation temperature. The following formula is used to compensate the temperature influence. IHTRU"IRJFX*(1+ȼ*(9RJFX-9HTRU ))
Temperature compensated specific gravity
8," IHTRU(8FRUQJ) IHTRU(-24)
Specific gravity at the measurement temperature
Temperature compensated density

See also Define the temperature-compensation coefficient α } Page 29 Configure the result conversion } Page 24 Density of pure water (0…40 °C) } Page 68 Change the density unit } Page 18
Available calculations:
See also Configure the result conversion } Page 24
The conversion of the measured value to the selected unit is based on the tables published by the American Petroleum institute (API). The following tables are available.
The Compensation temp. parameter defines the compensation temperature.
The Output format defines to which of the following units the result is converted:
See also Configure the result conversion } Page 24
The temperature influence is compensated by the temperature coefficient α as defined in Alpha*1000 > Alpha entry.

See also Define the temperature-compensation coefficient α } Page 29 Configure the result conversion } Page 24
Available conversions:
See also Configure the result conversion } Page 24
Available conversions:
See also Configure the result conversion } Page 24
You can define a function to convert a density or specific gravity to a concentration. The following steps are involved:
Define the density or specific gravity } Page 27 Define the function } Page 28 Define the format of the result } Page 29 Configure the result conversion } Page 24
4.9.6.7.1 Define the density or specific gravity
The conversions are either based on temperature compensated densities (d comp.) or specific gravities (SG).
Define a temperature-compensated density
The following formula is used to calculate the temperature-compensated density. IHTRU"IRJFX*(1+ȼ*(9RJFX-9HTRU ))
See also

2 Navigate to Compensation temp. and define the compensation temperature.
3 Navigate to Alpha*1000 > Alpha entry and define α. See [Define the temperature-compensation coefficient α } Page 29]
Define a temperature-compensated specific gravity The following formula is used to calculate the temperature-compensated specific gravity.
8," IHTRU(8FRUQJ) IHTRU(-24)
2 Navigate to Compensation temp. 1 and define the compensation temperature for the sample.
3 Navigate to Compensation temp. 2 and define the compensation temperature for water.
4 Navigate to Alpha*1000 > Alpha entry and define α. See [Define the temperature-compensation coefficient α } Page 29]
See also Define the function } Page 28
4.9.6.7.2 Define the function
You can either use a linear function (Linear (Ax+B)) or a second order polynomial function (2nd order pol. (ax²+bx+c). The functions are based on a graph with the density or specific gravity on the x-axis and the concentration on the y-axis. The following formulae are used.
3 Enter the values for the coefficients.
Calculate the coefficients A, B and C 1 If you want to use a second order polynomial function, prepare at least 4 solutions with different concentrations. 2 If you want to use a linear function, prepare at lest 2 solutions with different concentrations. 3 Determine the density or specific gravity you want to use for each concentration. 4 Plot the concentration (y-axis) against the density or specific gravity (x-axis). 5 Fit a line or second order polynomial to the curve.
Define the density or specific gravity } Page 27
See also

4.9.6.7.3 Define the format of the result
Using the parameters Unit and Decimal places, you can define the unit and the number of decimal places for the display of the current value and the result.
4.9.6.8 Define the temperature-compensation coefficient α
The temperature-compensation coefficient α is used to convert a measured value to a reference temperature. You have three options to define α.
See also Density and specific gravity } Page 25 Baumé scale } Page 26 Define the density or specific gravity } Page 27
4.9.6.8.1 Enter a known value for α
1 Navigate to Alpha*1000 > Alpha entry and select Manual.
3 Navigate to Alpha*1000. 4 Enter the value for α*1000.
1 Measure the density (d1) at a temperature (T1) above the normal measuring temperature. 2 Measure the density (d2) at a temperature (T2) below the normal measuring temperature. 3 Use the measured densities and the temperatures in [°C] to calculate α according to the following formula. 4 Calculate the result of α*1000.
Example Density at 26 °C (T1) = 0.7844 g/cm3
Density at 15 °C (T2) = 0.7937 g/cm3
See also Temperature-compensation coefficients α } Page 68
4.9.6.8.2 The density meter calculates a fixed value for α
2 Navigate to Density 1 and enter the value for d1. 3 Navigate to Temperature 1 and enter the value for T1.
Calculate α

4 Navigate to Density 2 and enter the value for d2. 5 Navigate to Temperature 2 and enter the value for T2. The density meter calculates the value for Alpha*1000 and displays it.
See also List of errors and problems } Page 57
4.9.6.8.3 Define the second order polynomial used to calculate α
The density meter can use the following second order polynomial to calculate α.
Procedure ▪ The three coefficients a, b and c are known. 1 Navigate to Alpha*1000 > Alpha entry and select 2nd order pol. (ax²+bx+c).
3 Enter the values for the coefficients a, b and c.
Calculate the coefficients a, b and c 1 Prepare at least 4 solutions with different concentrations. 2 Define two temperatures, one above und one below the normal measurement temperature. 3 Measure the density of each solution at both temperatures. 4 Calculate the temperature-compensation coefficient α for each solution. 5 Plot the values for α (y-axis) against the densities (x-axis) measured at one of the temperatures. 6 Fit a second order polynomial to the curve.
See also List of errors and problems } Page 57
NOTICE
Loss of data! A maximum of 1100 results is saved in the density meter. When the result list contains 1100 results and a new results is saved, the oldest result is overwritten. − Print or export the results before the maximum number of results is reached.
See also Change a method } Page 22 Change a test } Page 22 Change an adjustment } Page 22
You can define what kind of identification is assigned to a sample or to a standard.
Automatic numbering is only available for methods.
Sample ID Standard ID Sample ID entry Standard ID entry Action of the system
No identification is assigned to the sample or the standard.
ȼ"{F*IRJFX · None Not available

Sample ID Standard ID Sample ID entry Standard ID entry Action of the system
Users enter the identification before the density measurement starts.
An automatically generated identification is assigned to the sample. The identification automatically increases by 1, when a new measurement is started with the same method.
Automatic numbering Not available
The Printout parameter defines whether a result is automatically printed when it is saved and which infor- mation is printed.
Information Methods Test Adjustment
Date / Time Short/Long Short/Long Short/Long
Method ID Short/Long – –
Sample ID Short/Long – –
Application name with standard
– Short/Long Short/Long
Standard ID – Short/Long Short/Long
Status Short/Long Short/Long –
Calculation Short/Long – –
Density Short/Long Short/Long Short/Long
Dev. d – Short/Long Short/Long
Temp. Short/Long Short/Long Short/Long
d nominal – Short/Long Short/Long
SNR Long Long Long
Fixed 0…25 characters · Variable Not available

Information Methods Test Adjustment
Additional settings like the compensation temperature and the temperature-compen-
See also Connect a printer } Page 15
Results are automatically exported to EasyDirect Density & Refractometry if the density meter is connected to a computer with the software installed.
See also Connect to EasyDirect Density & Refractometry } Page 17
Result limits are only available for methods.
The Result limits parameter defines whether the density meter evaluates if a result lies within a defined range. The parameters Lower limit and Upper limit define the range.
1 Press the Menu key.
2 Navigate to Methods > Method settings > Calculation > Result limits.
3 To deactivate Result limits, navigate to No.
4 To activate Result limits, navigate to Yes.
Lower limit and Upper limit are displayed.
6 Set the Lower limit and Upper limit.
Use water as standard ▪ The Test window or the Adjustment window are open. 1 Navigate to Standard type and select Water.
The density meter uses predefined values for the Test or the Adjustment.
Use another substance as standard ▪ The Test window or the Adjustment window are open. 1 Navigate to Standard type and select Other standard.
3 Navigate to Compensation temp. and enter the compensation temperature.
4 Navigate to d nominal and enter the nominal density of the standard at the temperature you entered as compensation temperature. 5 Navigate to Alpha*1000 and define the temperature-compensation coefficient α.
Long Long Long · sation coefficient α · Procedure

See also Define the temperature-compensation coefficient α } Page 29
The Tolerance d parameter is used to evaluate the measured density and decide if the test passed or failed.
Recommendations for defining the value for Tolerance d:
The operator repeatability is the standard deviation of ten density determinations performed by the operator.

5.1 Start up the density meter
− Press key 1 (1).

A short beep is played. A welcome message is displayed.
The home screen opens. The density meter is ready
to use.
− Press and hold any of the keys or the navigation joystick for more than 3 s.
A beep is played and the screen goes dark.
The control circuits for the keys and the navigation joystick are energized. The rest of the density meter is
no longer energized.
Shut down using the Menu ▪ The home screen is open. 1 Press the Menu key.
2 Navigate to Shut down.
The dialog Do you want to shut down the instrument? opens.
3 Press the Yes key. A beep is played and the screen goes dark.
The control circuits for the keys and the navigation joystick are energized. The rest of the density meter is
no longer energized.
See also Activate and deactivate a cleaning reminder } Page 20
34 Operation Eagle Eye Power Solutions - SG-Pro

1 Press the Menu key. 2 To navigate to a desired menu, move the navigation joystick up or down. 3 To navigate to a submenu of the selected menu, move the navigation joystick to the right. 4 To go back, move the navigation joystick to the left.

In every menu, where the function of key 3 is marked with a green bar, you can press the navigation joystick to confirm the selection.
Shortcut to the method list
The Methods window opens.
3 To confirm the entered information and leave the keyboard window, press the OK key.

Min. 1 Max. 25
1 5 2 3
No. Icon Function
Cycles between the screens with small, capital or special characters.
2 Deletes all entered characters.
3 Deletes the character to the left of the cursor.
Operation 35 Eagle Eye Power Solutions - SG-Pro

No. Icon Function
4 Moves the cursor.
See also Change the keyboard layout } Page 20
Density determinations include the following two phases.
Because the measuring cell always contains residue from the previous sample or the cleaning solution, it is important that you remove the residue using the new sample. This is only possible if the residue is soluble in the new sample.
To achieve good results, make sure that the measuring cell contains only the sample you want to measure and that there are no bubbles in the measuring cell.
You can use the sample pump or syringes to fill, drain and rinse the measuring cell. The sample pump is used for samples with a viscosity up to 100 mPa*s. Syringes are typically used if the viscosity of the samples is above 100 mPa*s.
See also Example: Density determination using the sample pump } Page 36 Example: Density determination using a syringe } Page 39 Example: Density determination of cold or hot samples } Page 42
At the end of this phase, the residue in the measuring cell must be soluble in the next sample. The solubility of the sample defines if you can drain the measuring cell or if you need to rinse it.
Drain If the next sample dissolves the sample you just measured, you can drain the cell.
If the next sample does not dissolve the sample you just measured, you have to rinse the measuring cell with a cleaning solution. The cleaning solution must dissolve the sample that you just measured and the next sample. When you rinse the measuring cell, the cleaning solution removes the residue from the original sample.
See also Clean the measuring cell } Page 50
The following chapters show you how to configure and perform a density determination of deionized water at room temperature (23 °C or 73.4 °F).
36 Operation Eagle Eye Power Solutions - SG-Pro
Rinse

Summary of the configuration
You can find more information about the configuration of a method in the chapter [Configure density deter- minations } Page 21].
The Methods window opens.
2 Press the New key.
The Method settings window opens.
3 Navigate to Method ID and enter a name with up to 25 characters.
The method is listed in the Methods window. 5 Change the parameter settings to the values shown in the following table.
Workflow Guided The density meter prompts users to fill and drain the measuring cell.
Sample ID entry Variable The density meter prompts users to enter the sample ID before the density is measured.
Fill cycle 3 The density meter prompts users to fill and drain the measuring cell twice and then fill the measuring cell for the measurement.
Fill speed High The pump fills the measuring cell with 80 % of the maximum pump speed.
Measurement reliability Medium The result is saved as soon as the measured temperature value stays within 0.4 °C (0.72 °F) for 10 s.
Calculation > Category Density A density is calculated.
d The density is not compensated to a certain temperature.
Calculation > Result limits
Yes The density meter evaluates if the result is within a defined range.
Lower limit 0.9972 g/cm3 Defines the lower limit of the range.
Upper limit 0.9977 g/cm3 Defines the upper limit of the range.
Clean Drain Users are prompted to drain the measuring cell.
Printout None The result is not printed.
See also Change the density unit } Page 18 Activate and deactivate the confirmation of results } Page 24 Configure density determinations } Page 21
Operation 37 Eagle Eye Power Solutions - SG-Pro
Parameter Setting Explanation · Calculation > Calcu- lation

3 Press the Start key.
The Sample ID window opens. 4 Enter the identification.
Rinse the measuring cell
1 Immerse the filling tube into the sample (1).

The sample pump starts and stops when the
piston is in the uppermost position. 3 Hold the filling tube over the waste beaker (2).

The sample pump starts and stops when the
piston is in the lowermost position. 5 Clean the end of the filling tube with a clean tissue. 6 Repeat the steps once.
Fill the measuring cell and measure the density
1 Immerse the filling tube into the sample (1).
The sample pump starts and stops when the
piston is in the uppermost position. 3 Make sure there are no bubbles in the measuring cell. The measurement starts.

When the background of the measuring ribbon
changes color, the measurement is finished. The measured value is saved in the results.
38 Operation Eagle Eye Power Solutions - SG-Pro
Start the method

Drain the measuring cell
1 Hold the filling tube over the waste beaker (1).

The sample pump starts and stops when the
piston is in the lowermost position. 3 Clean the end of the filling tube with a clean tissue.
The following chapters show you how to configure and perform a density determination of a viscous sample such as 40 % w/w sugar solution at room temperature (23 °C or 73.4 °F).
Summary of the configuration
You can find more information about the configuration of a method in the chapter [Configure density deter- minations } Page 21].
The Methods window opens.
2 Press the New key.
The Method settings window opens.
3 Navigate to Method ID and enter a name with up to 25 characters.
The method is listed in the Methods window. 5 Change the parameter settings to the values shown in the following table.
Workflow Guided The density meter prompts users to fill and drain the measuring cell.
Sample ID entry Variable The density meter prompts users to enter the sample ID before the density is measured.
Fill cycle 2 The density meter prompts users to rinse the measuring cell and then fill the measuring cell for the measurement.
Fill speed High Has no effect because the syringe mode is activated.
Measurement reliability Medium The result is saved as soon as the measured temperature value stays within 0.4 °C (0.72 °F) for 10 s.
Operation 39 Eagle Eye Power Solutions - SG-Pro
Parameter Setting Explanation

Calculation > Category Sugar A sugar concentration is calculated.
Brix [%w/w] A Brix value is calculated.
Calculation > Result limits
None The density meter does not evaluate if the result is within a defined range.
Clean Drain Users are prompted to drain the measuring cell.
Printout None The result is not printed.
Activate and deactivate the confirmation of results } Page 24
2 Navigate to the method you want to use and press the Start key. The home screen opens.
3 Press the Start key.
The Sample ID window opens. 4 Enter the identification.
Rinse the measuring cell
1 Fill a new syringe with sample. 2 Insert the syringe into the connector for the syringe.

40 Operation Eagle Eye Power Solutions - SG-Pro

3 Hold the filling tube over the waste beaker (1). 4 Press the syringe plunger continuously into the syringe.

The sample flows through the measuring cell
and removes the residue from the previous sample. 5 Pull the plunger continuously out of the syringe. The sample is pulled back into the syringe and
the measuring cell fills with air. 6 Remove the syringe and empty it into a suitable waste beaker. 1
Fill the measuring cell and measure the density
1 Fill a new syringe with sample. 2 Insert the syringe into the connector for the syringe.

3 Hold the filling tube over the waste beaker (1). 4 Press the syringe plunger slowly (5…10 cm/s) and continuously into the syringe until only a small amount of sample is left in the syringe.

The sample flows with a laminar flow into the
measuring cell. 5 Leave the syringe in the connector. 6 Make sure there are no bubbles in the measuring cell.
The measurement starts. 1 When the background of the measuring ribbon
changes color, the measurement is finished. The measured value is saved in the results.
Operation 41 Eagle Eye Power Solutions - SG-Pro

Drain the measuring cell
1 Pull the plunger continuously out of the syringe. The sample is pulled back into the syringe and

the measuring cell fills with air. 2 Remove the syringe and empty it into a suitable waste beaker.
To achieve good results, the measuring cell and the sample must have the same temperature. If you fill the measuring cell with a sample that is more than 5 °C (9 °F) above or below the temperature displayed on the home screen, this condition is not met. Because the density meter has no active temperature control, you must wait until the temperature of the measuring cell and the temperature of the sample are the same, before you make the measurement. The following chapters show you how to configure and perform a density determination of deionized water that is more than 5 °C (9 °F) above or below the temperature displayed on the home screen.
Summary of the configuration
You can find more information about the configuration of a method in the chapter [Configure density deter- minations } Page 21].
The Methods window opens.
2 Press the New key.
The Method settings window opens.
3 Navigate to Method ID and enter a name with up to 25 characters.
The method is listed in the Methods window.
5 Change the parameter settings to the values shown in the following table.
Workflow Manual Users define their individual measuring workflows.
Sample ID entry Variable Users need to enter the identification before they start the measurement.
Fill speed High The pump fills the measuring cell with 80 % of the maximum pump speed.
Measurement reliability Medium The result is saved as soon as the measured temperature value stays within 0.4 °C (0.72 °F) for 10 s.
42 Operation Eagle Eye Power Solutions - SG-Pro
Parameter Setting Explanation

Calculation > Category Density A density is calculated.
d The density is not compensated to a certain temperature.
Calculation > Result limits
None The density meter does not evaluate if the result is within a defined range.
Clean Drain Users are prompted to drain the measuring cell.
Printout None The result is not printed.
Activate and deactivate the confirmation of results } Page 24
2 Navigate to the method you want to use and press the Start key. The home screen opens.
3 To enter a Sample ID, move the navigation joystick to the right.
The Sample ID window opens. 4 Enter the identification.
Rinse the measuring cell
1 Immerse the filling tube into the sample (1). 2 Move the navigation joystick upwards and let go.

The sample pump starts and stops when the
piston is in the uppermost position. 3 Hold the filling tube over the waste beaker (2). 4 Move the navigation joystick downwards and let go.

The sample pump starts and stops when the
piston is in the lowermost position. 5 Clean the end of the filling tube with a clean tissue. 6 Repeat the steps once.
Operation 43 Eagle Eye Power Solutions - SG-Pro

Fill the measuring cell and measure the density
1 Immerse the filling tube into the sample (1). 2 Move the navigation joystick upwards and let go.
The sample pump starts and stops when the
piston is in the uppermost position. 3 Make sure there are no bubbles in the measuring cell.

4 Press the Start key. The measurement starts.
When the background of the measuring ribbon
changes color, the measurement is finished. The measured value is saved in the results.
5 Press the OK key. 6 Wait for about 20 seconds and press the Start key 1 again. The sample that is still in the measuring cell is
measured again.
7 Tap Results and compare the last two results.
8 Tap Back. 9 If the results are not the same, continue to measure the same sample until the temperatures and the results of two measurements in a row are the same.
Drain the measuring cell
1 Hold the filling tube over the waste beaker (1). 2 Move the navigation joystick downwards and let go.

The sample pump starts and stops when the
piston is in the lowermost position. 3 Clean the end of the filling tube with a clean tissue.
With the RFID reader, you can read or write identifications for samples or standards to Smart Tags. The RFID reader only recognizes the following characters.
Metal containers interfere with reading and writing to Smart Tags. For this reason you need to place the Smart Tag (1) on a SmartSample film (2).
44 Operation Eagle Eye Power Solutions - SG-Pro

See also Miscellaneous accessories } Page 66
2 Navigate to Settings > RFID / Barcode >Write ID to RFID tag. The icon is displayed in the status ribbon. 3 Enter the identification. 4 Hold the RFID reader over a Smart Tag and press the Write key. An audio signal is played and the notification

RFID tag is written is displayed. The identification is saved on the Smart Tag.
Activate and deactivate the RFID reader (DensitoPro
only) } Page 13

press the RFID key and hold the RFID reader over the Smart Tag. An audio signal is played and the identification is
saved.
Read information from a Smart Tag ▪ The RFID reader is activated. 1 Press the Menu key.
2 Navigate to Settings > RFID / Barcode > Scan RFID tag.
The notification Scan RFID tag is displayed. 3 Hold the RFID reader over a Smart Tag and wait until an audio signal is played. The read information is displayed.

Activate and deactivate the RFID reader (DensitoPro
only) } Page 13 Configure density determinations } Page 21
With the barcode reader, you can read identifications for samples and standards from linear barcodes.
Operation 45 Eagle Eye Power Solutions - SG-Pro

The barcode reader only recognizes the following characters.
− When you are prompted to enter an identification, press the Barcode key and hold the barcode reader
over the barcode. An audio signal is played and the identification is saved.
See also Activate and deactivate the barcode reader (DensitoPro only) } Page 13 Configure density determinations } Page 21
You can only delete all results but not individual results or a subgroup of results. ▪ The home screen is open. 1 Press the Results key.
2 Press the Delete key.
The message Do you want to delete all results? is displayed.
3 Press the Yes key. An empty list is displayed.
NOTICE
Loss of data! A maximum of 1100 results is saved in the density meter. When the result list contains 1100 results and a new results is saved, the oldest result is overwritten. − Print or export the results before the maximum number of results is reached.
You can export or print results if you have the required accessories.
46 Operation Eagle Eye Power Solutions - SG-Pro

See also Accessories } Page 65 Connect to EasyDirect Density & Refractometry } Page 17 Connect a printer } Page 15
2 Navigate to Option and select EasyDirect.
4 Press the Start key. The export is completed and the result is displayed.
2 Navigate to Option and select EasyDirect.
4 To export all results, deactivate the filter. 5 To export a subset of results, configure the filter.
6 Press the Start key. The export is completed and the result list is displayed.
See also Configure the filter } Page 48 Connect to EasyDirect Density & Refractometry } Page 17
The Printout parameter defines whether a result is automatically printed when it is saved and which infor- mation is printed.
Information Methods Test Adjustment
Date / Time Short/Long Short/Long Short/Long
Method ID Short/Long – –
Sample ID Short/Long – –
Application name with standard
– Short/Long Short/Long
Standard ID – Short/Long Short/Long
Status Short/Long Short/Long –
Calculation Short/Long – –
Density Short/Long Short/Long Short/Long
Dev. d – Short/Long Short/Long
Temp. Short/Long Short/Long Short/Long
d nominal – Short/Long Short/Long
SNR Long Long Long
Operation 47 Eagle Eye Power Solutions - SG-Pro

Information Methods Test Adjustment
Additional settings like the compensation temperature and the temperature-compen-
2 Navigate to Option and select Print.
4 Navigate to Printout and set it to Short or Long as needed.
The result is printed.
2 Navigate to Option and select Print.
4 Navigate to Printout and set it to Short or Long as needed.
6 To print all results, deactivate the filter. 7 To print a subset of results, configure the filter.
8 Press the Start key.
The results are printed.
See also Configure the filter } Page 48 Connect a printer } Page 15
If you activate the filter, you can print or export a subset of the results. The following filters are available.
1 Press the Menu key.
2 Navigate to Results > Report > Filter.
3 Select the desired Filter type.
5 If you selected Range, define the values for Start date and End date.
See also Export results to EasyDirect Density & Refractometry } Page 47 Print results } Page 47
48 Operation Eagle Eye Power Solutions - SG-Pro
Long Long Long · sation coefficient α · Procedure

In this chapter you find descriptions of the maintenance tasks you should perform on your density meter. Any other maintenance tasks need to be performed by a service technician that has been qualified by EEPS.
If you experience problems with your density meter, contact your authorized EEPS dealer or service representative.
EEPS recommends that a preventive maintenance and calibration certification is done at least once a year through your authorized EEPS dealer or service representative.
If the standard operating procedures of your company require other maintenance intervals, use the intervals listed in the standard operating procedures.
Daily Clean the measuring cell at the end of the work day.
[Clean the measuring cell } Page 50]
Perform a test with deionized water. [Check the measurement accuracy } Page 51]
Activate and deactivate a cleaning reminder } Page 20
NOTICE
Danger of damage to the density meter due to inappropriate cleaning methods! Inappropriate cleaning agents can damage the housing or other parts of the density meter. If liquids enter the housing they can damage the density meter. 1 Make sure the cleaning agent is compatible with the material of the part you want to clean. 2 Make sure that no liquid enters the interior of the density meter.
If you have questions about the compatibility of cleaning agents, contact your authorized EEPS dealer or service representative.
EEPS recommends the following cleaning agents:
Procedure ▪ The density meter is shut down. −Wipe the housing with a cloth moistened with the cleaning agent.
Maintenance 49 Eagle Eye Power Solutions - SG-Pro
Frequency Task See · See also

Because the measuring cell always contains residue from the last sample, there is a risk that the measuring cell can be damaged by the sample. To prevent such damages, EEPS recommends that you clean the measuring cell at the end of each sample series.
To clean the measuring cell, you need to rinse the measuring cell with a cleaning solution. The cleaning solution must have the following properties.
Acids, concentrated Water (flush the measuring cell with plenty of water to remove the heat from the reaction of water and acid) Alkaline, concentrated 0.3…0.5 % deconex solution
Samples with fats or oily compoents 0.3…0.5 % deconex solution
Petrochemical samples Edible oils and fats
Toluene, xylene or petrol ether mixtures
Activate and deactivate a cleaning reminder } Page 20

The sample pump starts and stops when the
piston is in the uppermost position. 3 Lift the filling tube out of the cleaning solution and hold it over the waste beaker (2). 4 Move the navigation joystick downwards and let go.

The sample pump starts and stops when the
piston is in the lowermost position. 5 Clean the end of the filling tube with a clean tissue. 6 Repeat the steps until all residue from the sample is removed. 7 Move the navigation joystick upwards and let go.
The sample pump fills the measuring cell with
air. 8 Move the navigation joystick downwards and let go. Most of the cleaning solution is removed from the measuring cell. 9 Clean the end of the filling tube with a clean tissue.
50 Maintenance Eagle Eye Power Solutions - SG-Pro


3 Hold the filling tube over the waste beaker (1). 4 Press the plunger continuously into the syringe until the syringe is empty.

5 Pull the plunger continuously out of the syringe.

The cleaning solution is pulled back into the
syringe and the measuring cell fills with air. 6 Remove the syringe and empty it into a suitable waste beaker. 7 Repeat the steps until all residue from the sample is removed.
To check the measurement accuracy of the density meter, you need to configure and perform a test.
You can find more information about the configuration of a test in the chapter [Configure density determi- nations } Page 21].
The following chapters show you how to configure and perform a guided test with deionized water at room temperature (23 °C or 73.4 °C).
Maintenance 51 Eagle Eye Power Solutions - SG-Pro

2 Navigate to Test.
The Test window opens.
3 Change the parameter settings to the values shown in the following table.
Workflow Guided The density meter prompts users to fill and drain the measuring cell.
Standard type Water The density meter uses predefined values for the water measurement.
Standard ID entry Variable The density meter prompts users to enter the identifi- cation before the density is measured.
Fill cycle 3 The density meter prompts users to fill and drain the measuring cell twice and then fill the measuring cell for the measurement.
Fill speed High The pump fills the measuring cell with 80 % of the maximum pump speed.
Tolerance d 0.0002 g/cm3 The result is marked as passed if it falls within the following range.
d nominal ± Tolerance d
Clean Drain Users are prompted to drain the measuring cell.
Printout None The result is not printed.
See also Activate and deactivate the confirmation of results } Page 24 Change the density unit } Page 18
2 Navigate to Test.
The Test window opens.
3 Press the Start key. The home screen opens.
4 Press the Start key.
The Standard ID window opens.
5 Enter the identification.
Rinse the measuring cell
1 Immerse the filling tube in the water. 2 Press the OK key. The sample pump starts and stops when the piston is in the uppermost position.
52 Maintenance Eagle Eye Power Solutions - SG-Pro
Parameter Setting Explanation · Start the test

3 Hold the filling tube over the waste beaker.
4 Press the OK key. The sample pump starts and stops when the piston is in the lowermost position. 5 Clean the end of the filling tube with a clean tissue. 6 Repeat the steps once.
Fill the measuring cell and measure the density 1 Immerse the filling tube in the water.
2 Press the OK key. The sample pump starts and stops when the piston is in the uppermost position. 3 Make sure there are no bubbles in the measuring cell. The measurement starts. When the background of the measuring ribbon changes color, the measurement is finished. The measured value is saved in the results.
Drain the measuring cell
1 Hold the filling tube over the waste beaker.
2 Press the OK key. The sample pump starts and stops when the piston is in the lowermost position. 3 Clean the end of the filling tube with a clean tissue.
1 Check if the correct standard has been used, and if needed repeat the test with the correct standard. 2 If the test continues to fail, clean the measuring cell with a cleaning solution that dissolves your samples and dissolves in the standard. 3 Repeat the test. 4 If the test continues to fail, repeat the test twice and compare all three results. 5 If the results are not the same, clean the measuring cell and perform tests until the test passes or the results of three consecutive tests are the same. 6 If the test continues to fail and the results of three consecutive tests are the same, restore the factory adjustment data and repeat the test. 7 If the test continues to fail, perform an adjustment.
See also Clean the measuring cell } Page 50 Restore the factory adjustment data } Page 58 Adjust the measurement accuracy } Page 58
The density meter works with a rechargeable 2400 mAh Li-Ion battery that you can order from EEPS.
Maintenance 53 Eagle Eye Power Solutions - SG-Pro

1 Press on the lower part (1) of the battery cover and push the battery cover up. 2 Pull the upper part of the battery out of the battery compartment and lift the battery out.


3 Insert the lower part of the battery at an angle into the battery compartment. 4 Push the upper part of the battery into the battery compartment. 5 Slide the battery cover from above onto the density meter.


See also Accessories } Page 65 Technical data } Page 62
1 Unscrew the filling tube by hand and remove it. 2 Install the new filling tube.

See also Accessories } Page 65 Install the filling tube } Page 13
Remove the sample pump
54 Maintenance Eagle Eye Power Solutions - SG-Pro

3 Press on the grip recesses (4) of the cover and lift the lower part of the cover out of the housing. 4 Pull the top of the cover out of the housing. 5 Lift the piston (1) out of the guide system. 6 Pull the cylinder (2) up and out of its guide system (3).


Install the sample pump
1 Insert the cylinder (3) of the sample pump in its guide system (4). 2 Insert the piston (1) in its guide system (2). 3 To check if the sample pump works, move the navigation joystick downwards and let go.

The sample pump starts and stops when the
piston is in the lowermost position. 4 Reinstall the cover of the sample pump.
See also Accessories } Page 65
It is possible to update the firmware of your density meter. To update the firmware of the density meter, contact an authorized EEPS dealer or service representative.
The density meter can be stored at temperatures of -20…+50 °C (-4…+122 °F). But the lifespan of the battery is reduced when it is stored at high temperatures. 1 Clean the measuring cell. 2 Shut down the density meter. 3 Clean the housing. 4 Store the density meter in a dry and clean place.
See also Clean the density meter } Page 49 Technical data } Page 62
If you have questions about transporting your density meter, contact your authorized EEPS dealer or service representative.
1 Shut down the density meter.
Maintenance 55 Eagle Eye Power Solutions - SG-Pro

2 Clean the density meter. 3 Ship the density meter using the original packaging.
In conformance with the European Directive 2012/19/EU on Waste Electrical and Electronic Equipment (WEEE) this device may not be disposed of in domestic waste. This also applies to countries outside the EU, per their specific requirements.
Please dispose of this product in accordance with local regulations at the collecting point specified for electrical and electronic equipment. If you have any questions, please contact the responsible authority or the distributor from which you purchased this device. Should this device be passed on to other parties (for private or professional use), the content of this regulation must also be related.
Thank you for your contribution to environmental protection.
56 Maintenance Eagle Eye Power Solutions - SG-Pro

7.1 List of errors and problems
Problem Possible cause Measure
The sample pump does not aspirate any liquid.
The locking screw is not installed. − Install the locking screw.
The locking screw is not tight. − Tighten the locking screw. The measuring cell is not clean. See [Measures if a test fails } Page 53].
Test of measurement accuracy fails.
Mismatch between measured and configured standard
See [Measures if a test fails } Page 53].
The tolerance range is too small. See [Configure the tolerance for a
test } Page 33].
The density meter is not adjusted correctly.
See [Measures if a test fails } Page 53]. 1 Hold the filling tube over the waste beaker. 2 Move the navigation joystick downwards and let go. The sample pump starts
The message Drain cell over waste beaker. is displayed when you try to start a method.
The piston is not in the lowermost position.
and stops when the piston is in the lowermost position. 3 Clean the end of the filling tube with a clean tissue.
There are bubbles in the measuring cell.
Fill speed is too high. 1 If you use the sample pump, lower the setting for fill speed. See [Define the pump speed } Page 23]. 2 If you use a syringe, press the syringe plunger more slowly into the syringe.
The liquid contains dissolved gases.
− Degas the sample before you
fill it into the measuring cell.
Connector for syringe or the filling tube is not tight.
1 Tighten the connector of the filling tube. 2 Tighten the locking screw or the connector for the syringe.
Troubleshooting 57 Eagle Eye Power Solutions - SG-Pro

Problem Possible cause Measure
The message Pump position unknown. is displayed.
The saved pump position was lost because the battery was removed too long.
The density meter is set up to work with a syringe.
− You can ignore the message
and continue to work with the density meter.
The density meter is set up to work with the sample pump. 1 Hold the filling tube over the waste beaker. 2 Move the navigation joystick downwards and let go. The sample pump starts
and stops when the piston is in the lowermost position. 3 Clean the end of the filling tube with a clean tissue.
"--" is displayed, instead of a value for Alpha*1000.
The calculated value is out of range.
− Check and correct the density
values and the values for the compensation temperatures you entered.
"--" is displayed, instead of a The measured value or the result is 1 Check that the unit defined in measured value or a result. out of range. the method matches the sample you want to measure. 2 If you use a user-defined conversion, check that all the settings are correct.
If you restore the factory adjustment data, the current adjustment data is overwritten with the adjustment data that were determined when the density meter was adjusted during production.
1 Press the Menu key.
2 Navigate to Maintenance & Service > Reset to factory adjust..
The message Do you want to reset to factory adjustment? opens.
3 To overwrite the current adjustment data with the factory adjustment data, press Yes.
The factory adjustment data are restored.
NOTICE
The following chapters show you, how to configure and perform an exemplary adjustment with water.
You can find more information about the configuration of an adjustment in the chapter [Configure density determinations } Page 21].
58 Troubleshooting Eagle Eye Power Solutions - SG-Pro

See also Measures if a test fails } Page 53
The following chapters show you how to configure and perform a guided adjustment with deionized water at room temperature (23 °C or 73.4 °C).
2 Navigate to Adjustment.
The Adjustment window opens.
3 Change the parameter settings to the values shown in the following table.
Workflow Guided The density meter prompts users to fill and drain the measuring cell. Standard type Water The density meter uses predefined values for the water measurement.
Standard ID entry Variable The density meter prompts users to enter the identifi- cation before the density is measured. Fill cycle 3 The density meter prompts users to fill and drain the measuring cell twice and then fill the measuring cell for the measurement.
Fill speed High The pump fills the measuring cell with 80 % of the maximum pump speed. Clean Drain Users are prompted to drain the measuring cell.
Printout None The result is not printed.
2 Navigate to Adjustment.
The Adjustment window opens.
3 Press the Start key. The home screen opens.
4 Press the Start key.
The Standard ID window opens.
5 Enter the identification.
Troubleshooting 59 Eagle Eye Power Solutions - SG-Pro
Parameter Setting Explanation · Start the adjustment

Rinse the measuring cell
1 Immerse the filling tube in the water (1).

The sample pump starts and stops when the
piston is in the uppermost position. 3 Hold the filling tube over the waste beaker (2).

The sample pump starts and stops when the
piston is in the lowermost position. 5 Clean the end of the filling tube with a clean tissue. 6 Repeat the steps once.
Fill the measuring cell and measure the density 1 Immerse the filling tube in the water (1).
The sample pump starts and stops when the
piston is in the uppermost position. 3 Make sure there are no bubbles in the measuring cell. The measurement starts.

When the background of the measuring ribbon
changes color, the measurement is finished. The measured value is saved in the results. 4 Accept or reject the adjustment.
Drain the measuring cell
1 Hold the filling tube over the waste beaker (1).

The sample pump starts and stops when the
piston is in the lowermost position. 3 Clean the end of the filling tube with a clean tissue.
Activate and deactivate the confirmation of
results } Page 24
60 Troubleshooting Eagle Eye Power Solutions - SG-Pro
See also

NOTICE
Loss of data With a factory reset all user-specific settings are set to the default values and all saved data (e.g. sample IDs, results, Method IDs and adjustment data) are deleted.
−Document the data you need before you restore the factory settings.
1 Press the Menu key.
2 Navigate to Maintenance & Service > Reset to factory settings.
The dialog Do you want to reset all values to factory settings? opens.
3 To reset the density meter to factory settings, press the Yes key.
The density meter restarts.
2 Navigate to Information.
The firmware version and other system information are displayed.
Instrument type Shows if the density meter is a SG-Pro or DensitoPro.
SNR instrument Shows the serial number of the instrument.
Firmware version Shows the firmware version of the density meter. Licenses Opens a window with information on the End User License Agreement, third party licenses and open source attribution files.
Troubleshooting 61 Eagle Eye Power Solutions - SG-Pro

8.1 Density meter
Power rating instrument Input values 5 V DC, 1.0 A
Connector type USB C Power rating AC adapter Input values 100…240 V AC, 0.3 A
Input frequency 50 - 60 Hz Output values 5 V DC, 1.0 A
Battery Type Rechargeable lithium-ion battery Capacity 2400 mAh Temperature range for charging 0…40 °C Dimensions Width 56 mm
Length 288 mm Depth 80 mm
Weight 355 g Screen Technology Color screen Size 2.4"
Resolution 240 x 320 pixels USB type A Data rate USB 2.0, host fullspeed
Output voltage 5 V DC ±5 % Max. output current 250 mA
USB type C Data rate USB 2.0, device fullspeed
Input voltage 5 V DC ±10 % Max. input current 400 mA, max. 1 A with AC adapter
RFID reader (DensitoPro only)
Compatibility METTLER TOLEDO Smart Tags
Barcode reader (DensitoPro only)
Materials Housing PBT (polybutylene terephthalate) Measuring cell Borosilicate glass
Parts in contact when working with sample pump
PTFE (polytetrafluorethylene) PVDF (polyvinylidene difluoride) EPDM (ethylene propylene diene monomer (M- class) rubber) FFKM (perfluoro-elastomers) Borosilicate glass Filling tube PTFE (polytetrafluorethylene)
Connector for filling tube or syringe
Ambient temperature (SG-Pro) -10…+50 °C Ambient temperature (DensitoPro) 0…+50 °C
Relative humidity Max. 80 % (non-condensing) at 31 °C, linear fall to 50 % at 40 °C
62 Technical data Eagle Eye Power Solutions - SG-Pro

Altitude Up to 2000 m above sea level
Use In interior spaces
Instrument -20 °C…+50 °C
Battery Storage for up to 1 month -20 °C…+60 °C
Storage for up to 3 months -20 °C…+45 °C
Storage for up to 1 year -20 °C…+30 °C
Directives, standards and REACH regulation
Directives and standards complied with are listed on the declaration of conformity. SVHC candidate substances according to REACH (Article 33)
Material CAS No.
PZT (Lead Zirconate Titanate) 12626-81-2
Density Measuring range 0.000…3.000 g/cm3
Accuracy 1) ± 0.001 g/cm3
Repeatability ± 0.0005 g/cm3 Resolution 0.0001 g/cm3
Temperature Temperature range for sample 0…50 °C
Resolution 0.1 °C
Accuracy ±0.2 °C Viscosity Using sample pump 0...100 mPa*s Using syringe 0...1000 mPa*s
Volumes Sample pump 5 mL Min. sample volume 2 mL
Temperature compensation Automatic or by using a user-defined temperature correction coefficient
Application Max. number of methods 30
Ethanol (d) % w/w Range 0.0. 100.0 % w/w
Accuracy ±1.0 % w/w Ethanol (d) % v/v Range 0.0 % v/v
Accuracy ±1.0 % v/v
Technical data 63 Eagle Eye Power Solutions - SG-Pro

Proof (IP) Range 0.0. 175.0 Proof
Accuracy ±1.75 Proof
Proof (US) Range 0.0. 200.0 Proof
Accuracy ±2.0 Proof
Light Baumé Range 10...100 °Bé
Accuracy ±0.4 °Bé
Heavy Baumé Range 0...72 °Bé
Accuracy ±0.1 °Bé
H2SO4 % w/w Temperature range 10…50 °C
Concentration range 0.0. 100.0 % w/w
Accuracy ±1.0 % w/w
H2SO4 % v/v Temperature range 10…50 °C
Concentration range 0.0 % v/v
Accuracy ±1.0 % v/v
Brix (d) % w/w Range 0.0. 85.0 % w/w
Accuracy ±0.3 % w/w
Plato Range 0.0. 85.0 °P
Accuracy ±0.3 °P
International standards and norms
International standards and norms complied with are listed on the internet.
64 Technical data Eagle Eye Power Solutions - SG-Pro
Characteristic Value

The following accessories and parts are available for the density meters SG-Pro and DensitoPro.
All parts are specified with their order number and quantity in cases where more than one part is included. When ordering, some parts are only available in a minimum order quantity. In such cases, the corre- sponding minimum order quantity is quoted. Certain parts are only available in specific countries. If you have any questions, contact your authorized EEPS dealer.
Filling Tube 190 mm 30330847

Filling tube 600 mm 30330848
Locking Screw, SG-Pro 30330852
Connector syringe/filling tube 30330853

Disposable syringe (100 pcs) 10 mL 51338100

Sample pump cover 30330851
Accessories 65 Eagle Eye Power Solutions - SG-Pro
Sample pump · 30330849

Bluetooth Printer Godex MX20 with Dongle 30330864
Bluetooth Printer Paper (5 rolls) 30330865
Bluetooth Dongle Godex MX20 30330863

Lab equip access. data writer USB-P25/01 11124301
EasyDirect Density & Refractometry PC Software 30451628
Density standards (10 pcs) 6 mL 51325005
Battery Li-Ion 2400mAh 30330855

66 Accessories Eagle Eye Power Solutions - SG-Pro
Carrying Case 30330861


Power Supply and Worldwide Adapters 30449255

Tag SmartSample (10 pcs) 30449268
Tag SmartSample Film for Tanks (10 pcs) 30449269
Accessories 67 Eagle Eye Power Solutions - SG-Pro
Battery cover 30330854 · USB-C Cable 30449253 · Accessories for DensitoPro

10.1 Density of pure water (0…40 °C)
Temperature [°C] Density [g/cm3] Temperature [°C] Density [g/cm3]
0 0.99984
1 0.99990 21 0.99799
2 0.99994 22 0.99777
3 0.99996 23 0.99754
4 0.99997 24 0.99730
5 0.99996 25 0.99705
6 0.99994 26 0.99679
7 0.99990 27 0.99652
8 0.99985 28 0.99624
9 0.99978 29 0.99595
10 0.99970 30 0.99565
11 0.99961 31 0.99534
12 0.99950 32 0.99503
13 0.99938 33 0.99471
14 0.99925 34 0.99438
15 0.99910 35 0.99404
16 0.99894 36 0.99369
17 0.99878 37 0.99333
18 0.99860 38 0.99297
19 0.99841 39 0.99260
20 0.99821 40 0.99222
[Chemical Handbook Fundamental Version, Rev. 3, Table 5.2 (1984)]
Substance Temperature range [°C] α • 103/°C
Water 15…30 0.23
Ethanol 0…30 1.09
m-xylene 0…30 0.99
p-xylene 15…30 1.02
Glycerine 15…30 0.49
Chloroform 0…30 1.26
Carbon tetrachloride 0…30 1.22
Toluene 0…30 1.07
Benzene 6…30 1.21
Methanol 6…30 1.18
Acetone 0…30 1.42
Bromobenzene 0…30 0.91
Cyclohexane 0…30 1.20
Isopropanol 0…30 1.06
n-nonane 0…30 1.08
68 Appendix Eagle Eye Power Solutions - SG-Pro

Activate 20 Deactivate 20
Assemble 12 Change prongs 12 Connect 12 Accessories 65
Deactivate 13 Read 45
Auto rotation 20 Barcode reader 13 Cleaning reminder 20
Install 53 Remove 53 Replace 53
Confirmation of results 24 Password protection 19
Storage temperature 63 Technical data 62 Baumé 24, 26 Beep
RFID reader 13 Sample pump mode 13 Syringe mode 14
Configure 23, 29, 30, 31, 32 Configure example 59
Change 18 Dimming 18 Brix 27 C Mass percent 26
Perform 58 Perform example 59
Proof (IP) 26 Proof (US) 26
Volume percent 26 Alpha
Table 68 Ambient conditions 62
Density unit 18 Keyboard layout 20 Language 20 Method 22 Password 20 Prongs AC adapter 12 Screen brightness 18
American Petroleum Institute 24
API degree 26 Density 26 Specific gravity 26
AC adapter 12 Filling tube 13 Audio signal
Time 17 Time format 17
Density meter 12 Clean
Index 69 Eagle Eye Power Solutions - SG-Pro

Measuring cell 50 Cleaning reminder
Activate 20 Deactivate 20 Concentration
Sugar 24 Sulfuric acid 24 User-defined 24, 27
Auto rotation 20 Barcode reader 13 Cleaning reminder 20
Adjustment 22, 23, 29, 30, 31, 32
Confirmation of results 24 Password protection 19
Alpha 29 Drain 23
RFID reader 13 Sample pump 14 Sample pump mode 14 Syringe mode 13
Measurement reliability 24 Method 22, 23, 24, 29, 30, 31, 32 Result limits 32 Rinse 23 Standard 32
Method 22 Results 46 Density 24, 25
Test 22, 23, 29, 30, 31, 32, 33 Tolerance 33 User-guidance 23
Temperature compensated 25 Density determination
Adjustment 59 Guided 36, 39 Sample pump 36, 42 Syringe 39
Test 52 Unguided 42
Overview 8 Ship 55 Shut down 34
Store 55 Unpack 11
Density of pure water
Bluetooth printer 15, 16 EasyDirect Density & Refractometry 17 Godex MX20 15, 16 Power supply 12 USB printer 15, 16
Adjustment 18 Change 18
Automatic 18 Disposal 56 Drain
Install 15 Material 62
Background 36 Configure 23 Guided 39, 42 Sampel pump 44
70 Index Eagle Eye Power Solutions - SG-Pro

Syringe 42 Unguided 44
E
Clean 49 Material 62
EasyDirect Density & Refractometry
Connect 17 Export 32 End user license agreement 5 Energy saver 18
Number 35 Special character 35 Symbol 35 Text 35
Connector for syringe 15 Filling tube 13 Sample pump 55 Syringe adapter 15
F
Key functions ribbon 10 Keyboard
Change layout 20 Overview 35
Restore 61 Fahrenheit 18 Fill
Guided 38, 41 Sample pump 44 Syringe 41 Unguided 44
Fill speed 23 Filling tube
Connector for filling tube or syringe 62 Filling tube 62 Housing 62 Measuring cell 62 Sample pump 62 Measurement accuracy
Install 13 Material 62 Remove 54 Replace 54
Adjust 58 Check 51 Measurement reliability
End user license agreement 5 Open source attribution files 5 Third party licenses 5
Configure 24 Measuring cell 9 Clean 50 Material 62 Measuring ribbon 10 Method
Update 55 Version 5, 61
Index 71 Eagle Eye Power Solutions - SG-Pro

Configure 23, 24, 29, 30, 31, 32
Sample pump 54 Replace
Delete 22 Method ribbon 10
Battery 53 Filling tube 54 Sample pump 54
Factory adjustment 58 Factory settings 61
Open source attribution files 5
Factory adjustment 58 Result limits
Functions 8 Instrument 8 Keyboard 35
View 46 RFID reader
Change 20 Password protection 19 Activate 19 Deactivate 19
Adjustment 58 Test 51 Perform example
Adjustment 59 Method 38, 40, 43 Test 52 Plato 27
Barcode reader 45 RFID reader 45 Remove
Battery 53 Filling tube 54 Export 32, 47
72 Index Eagle Eye Power Solutions - SG-Pro

Metal container 44 Read 45 Write 45 Rinse
Configure 23 Guided 38, 40 Sample pump 43 Syringe 40 Unguided 43
Signal words 6 Warning symbols 6 Sample
Sample ID ribbon 10 Sample pump
Drain 44 Fill 44
Install 55 Material 62 Remove 54 Replace 54
Index 73 Eagle Eye Power Solutions - SG-Pro

Rinse 43 Set up 13 Sample pump mode
Density meter 55 Sucrose concentration 24 Sugar concentration 24
Sampling ribbon 10 Scope of delivery 11
Plato 27 Sulfuric acid concentration 24
Brightness 18 Dimming 18
Density meter 34 Switch on
Sample pump 13 Syringe 14 Ship
Density meter 34 Symbol 5, 10 Enter 35 Syringe
Density meter 55 Shortcut method list 35 Shut down
Automatic 18 Density meter 34 Signal words 6
Fill 41 Rinse 40
Set up 14 Syringe adapter
Site requirements 6, 62
Metal container 44 Read 45 Write 45 Special character
Specific gravity 24, 25 Temperature compensated 25 Standard
Configure 32 Identification 30 Standard delivery 11
Configure 29 Values 68
Sample pump 38 Syringe 40 Unguided 43
Configure 23, 29, 30, 31, 32, 33
Density unit 18 Failed 53
Perform 51 Perform example 52 Printer 16
74 Index Eagle Eye Power Solutions - SG-Pro

Enter 35 Third party licenses 5
List of errors 57
Temperature 18 Unpack 11
Connect 15, 16 User guidance
Index 75 Eagle Eye Power Solutions - SG-Pro