EN
Official user manual

Eagle Microsystems Handheld Density Meter (SG Pro)

77 pages·3 MB·Measuring Tool·PDF·ENDEFRITNLJA
Eagle Microsystems Handheld Density Meter (SG Pro)
Table of Contents 80 sections Reference Manual · Handheld Density Meter · SG-Pro · 1 Introduction · … ShowHide
Reference Manualp.1Handheld Density Meterp.1SG-Prop.11 Introductionp.72 Safety Informationp.82.1 Definitions of signal words and warning symbolsp.82.2 Product specific safety notesp.8WARNINGp.9NOTICEp.93 Design and Functionp.103.1 Instrument overviewp.103.2 Overview of functionsp.103.3 User interfacep.123.3.1 Home screenp.123.3.2 Icons on the screenp.124 Installation and commissioningp.134.1 Scope of deliveryp.134.2 Unpack the density meterp.134.3 Charge the density meterp.144.3.1 Charge with the AC adapterp.144.3.2 Charge with a computerp.144.4 Setup for working with the sample pumpp.154.4.1 Activate the sample pump modep.154.4.2 Install the filling tubep.154.5 Activate and deactivate the RFID reader (DensitoPro only)p.154.6 Activate and deactivate the barcode reader (DensitoPro only)p.154.7 Install accessoriesp.164.7.1 Setup for working with a syringep.164.7.1.1 Activate the syringe modep.164.7.1.2 Install the connector for the syringep.174.7.2 Connect a printerp.174.7.2.1 Connect a USB printerp.184.7.2.2 Connect a bluetooth printerp.184.7.2.3 Print a test pagep.184.7.3 Connect to EasyDirect Density & Refractometryp.194.8 Configure the density meterp.194.8.1 Change the date and time formatp.194.8.2 Change the temperature unitp.204.8.3 Change the density unitp.204.8.4 Change energy saver settingsp.204.8.5 Protect actions with a passwordp.214.8.5.1 Enable the password protectionp.214.8.5.2 Disable the password protectionp.214.8.5.3 Change the passwordp.224.8.6 Activate and deactivate the auto rotation of the screenp.224.8.7 Activate and deactivate the audio signalp.224.8.8 Change the keyboard layoutp.224.8.9 Change the languagep.224.8.10 Activate and deactivate a cleaning reminderp.224.9 Configure density determinationsp.234.9.1 Create, change or delete a methodp.234.9.1.1 Create a methodp.234.9.1.2 Change a methodp.244.9.1.3 Delete a methodp.244.9.2 Change a testp.244.9.3 Change an adjustmentp.244.9.4 Configure the user guidancep.254.9.4.1 Define the level of user guidancep.254.9.4.2 Define the pump speedp.254.9.4.3 Define the fill cycle (only guided workflow)p.254.9.4.4 Define the cleaning type (only guided workflow)p.254.9.4.5 Activate and deactivate the confirmation of resultsp.264.9.5 Configure the measurement reliabilityp.264.9.6 Configure the result conversionp.264.9.6.1 Density and specific gravityp.278,(Y/Y)" I (8FRUQJ) I ( - 2 4 )p.274.9.6.2 Alcohol concentrationp.284.9.6.3 American Petroleum Institute (API)p.284.9.6.4 Baumé scalep.284.9.6.5 Sulfuric acid concentrationp.294.9.6.6 Sugar concentrationp.294.9.6.7 User defined concentrationp.29ȼ"{ I 2 -1) / {9 -9 ) I 1p.312 +G*I RJFX +H)p.324.9.7 Configure the automatic documentation of resultsp.324.9.7.1 Configure the sample or standard identificationp.324.9.7.2 Configure the automatic printing of resultsp.334.9.7.3 Configure the automatic export of resultsp.344.9.8 Configure result limitsp.344.9.9 Configure a standardp.34
Page 1 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — Reference Manual
Page 1 · click to enlarge
Page 1
Eagle Microsystems Handheld Density Meter (SG Pro) — Reference Manual diagram (page 1)
Fig. 1 · Reference Manual

Reference Manual

Handheld Density Meter

SG-Pro

Eagle Microsystems Handheld Density Meter (SG Pro) — Reference Manual diagram (page 1)
Fig. 2 · Reference Manual
Eagle Microsystems Handheld Density Meter (SG Pro) — Reference Manual diagram (page 1)
Fig. 3 · Reference Manual
Page 2 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — Reference Manual
Page 2 · click to enlarge
Page 2
Page 2 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — Reference Manual
Page 3 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — Reference Manual
Page 3 · click to enlarge
Page 3
Page 3 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — Reference Manual
Page 4 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — Reference Manual
Page 4 · click to enlarge
Page 4
Page 4 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — Reference Manual
Page 5 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — Reference Manual
Page 5 · click to enlarge
Page 5
Page 5 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — Reference Manual
Page 6 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — Reference Manual
Page 6 · click to enlarge
Page 6

4 Table of Contents Eagle Eye Power Solutions – SG-Pro

Page 7 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 1 Introduction
Page 7 · click to enlarge
Page 7

1 Introduction

Thank you for choosing an Eagle Eye Power Solutions density meter. The handheld density meters SG-Pro and DensitoPro are handheld measuring instruments used to determine the density of liquids.

The instructions in this document refer to handheld density meters SG-Pro and DensitoPro running firmware version V1.0.0 or higher.

The firmware license is subject to the End User License Agreement EULA version 3.0.

If you have any additional questions, contact your authorized EEPS dealer or service represen- tative.

Refers to an external document.

Elements of instructions ▪ Prerequisites 1 Steps 2 ...

See also View the firmware version and other system information } Page 61

Introduction 5 Eagle Eye Power Solutions - SG-Pro

About this document · Conventions and symbols · Intermediate results Results

Page 8 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 2.2 Product specific safety notes
Page 8 · click to enlarge
Page 8

2 Safety Information

  • This Reference Manual contains a full description of the instrument and its use.
  • Keep the Reference Manual for future reference.
  • Include the Reference Manual if you transfer the instrument to other parties.

Only use the instrument according to the Reference Manual. If you do not use the instrument according to the Reference Manual or if it is modified, the safety of the instrument may be impaired and EEPS assumes no liability.

2.1 Definitions of signal words and warning symbols

Safety notes contain important information on safety issues. Ignoring the safety notes may lead to personal injury, damage to the instrument, malfunctions and false results. Safety notes are marked with the following signal words and warning symbols:

2.2 Product specific safety notes

The density meter is designed to be used by trained staff and is intended for measuring the density of liquid samples that are compatible with the materials with which they come into contact.

The density meter is designed for indoor and outdoor operation in dry conditions. The following site requirements apply:

  • The ambient conditions are within the limits specified in the technical data.
  • No corrosive gas atmosphere
  • No explosive atmosphere
  • No powerful electric or magnetic fields

Any other type of use and operation beyond the limits of use stated by EEPS without consent from EEPS is considered as not intended.

This device complies with Industry Canada licence-exempt RSS standard(s) and part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause interference, and (2) this device must accept any interference, including interference that may cause undesired operation of the device. Changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment.

Responsibilities of the instrument owner

The instrument owner is the person holding the legal title to the instrument and who uses the instrument or authorizes any person to use it, or the person who is deemed by law to be the operator of the instrument. The instrument owner is responsible for the safety of all users of the instrument and third parties.

EEPS assumes that the instrument owner trains users to safely use the instrument in their workplace and deal with potential hazards. EEPS assumes that the instrument owner provides the necessary protective gear.

6 Safety Information Eagle Eye Power Solutions - SG-Pro

Signal words · Electrical shock · Intended use

Page 9 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — NOTICE
Page 9 · click to enlarge
Page 9

WARNING

Danger of death or serious injury due to electric shock! Contact with parts that carry a live current can lead to death or injury. 1 Only use the AC adapter designed for your instrument. 2 Keep all electrical cables and connections away from liquids and moisture. 3 Check the cables and the plugs for damage and replace damaged cables and plugs.

NOTICE

Risk of damage to the instrument due to the use of unsuitable parts! Using unsuitable parts with the instrument can damage the instrument or cause it to malfunction.

−Only use parts from EEPS that are intended to be used with your instrument.

Safety Information 7 Eagle Eye Power Solutions - SG-Pro

Safety notes

Page 10 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 3.2 Overview of functions
Page 10 · click to enlarge
Page 10

3 Design and Function

3.1 Instrument overview

1 2 3 11 4 10

Eagle Microsystems Handheld Density Meter (SG Pro) — 3 Design and Function diagram (page 10)
Fig. 4 · 3 Design and Function
Eagle Microsystems Handheld Density Meter (SG Pro) — 3 Design and Function diagram (page 10)
Fig. 5 · 3 Design and Function

No. Name Function

1 USB-C socket To connect the density meter with the AC adapter or a computer.

2 RFID reader and barcode reader (DensitoPro only )

To scan a linear barcode or read and write information to Smart Tags.

3 Battery cover To access the battery.

4 USB-A socket To connect a USB printer or bluetooth dongle.

5 Sample pump cover To access the sample pump.

6 Sample pump window

To view the sample pump and to check if the sample pump is filled or empty.

7 Filling tube To fill and drain the measuring cell.

8 Filling opening with locking screw

To fill and drain the measuring cell using a syringe. The locking screw closes off the filling opening.

9 Measuring cell window

To view the measuring cell and to check if the measuring cell is filled or empty.

10 Navigation joystick To navigate through the menus.

11 Key 1 To start up the density meter, select menu items and to navigate.

12 Key 2 To select menu items and to navigate.

13 Key 3 To select menu items and to navigate.

14 Screen To display settings and results.

3.2 Overview of functions

The density meter uses the oscillating body method as measurement technology. The density meter has no active temperature control.

8 Design and Function Eagle Eye Power Solutions - SG-Pro

13 12 14

Page 11 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 3 Design and Function
Page 11 · click to enlarge
Page 11

If you measure samples with low viscosity, you can use the sample pump (1) and the filling tube (2) to fill the measuring cell (5). For samples with high viscosity you can use a syringe (3) and the connector for the syringe (4) to fill the measuring cell (5). The user can observe the filling through the window of the measuring cell.

Eagle Microsystems Handheld Density Meter (SG Pro) — 3 Design and Function diagram (page 11)
Fig. 6 · 3 Design and Function
Eagle Microsystems Handheld Density Meter (SG Pro) — 3 Design and Function diagram (page 11)
Fig. 7 · 3 Design and Function
Eagle Microsystems Handheld Density Meter (SG Pro) — 3 Design and Function diagram (page 11)
Fig. 8 · 3 Design and Function

For repeated measurements, you can define and save workflows as methods. Two workflow types are available:

  • Customizable guided workflow: guides the user step by step through the density determination.
  • Unguided workflow: the user decides which steps are needed.

The results are automatically converted into one of the following predefined units or a unit that the user defines.

  • Density
  • Specific gravity
  • Alcohol concentration
  • American Petroleum Institute (API density, API specific gravity, API degree)
  • Baumé scale
  • Sulfuric acid concentration
  • Sucrose concentration

Design and Function 9 Eagle Eye Power Solutions - SG-Pro

Page 12 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 3.3.2 Icons on the screen
Page 12 · click to enlarge
Page 12

3.3 User interface

3.3.1 Home screen

Eagle Microsystems Handheld Density Meter (SG Pro) — 3.3.1 Home screen diagram (page 12)
Fig. 9 · 3.3.1 Home screen

d [g/cm3]

Temp. 23 °C

No. Name Explanation

1 Status ribbon Displays the current date, time, battery charge status and connection to acces- sories like EasyDirect Density & Refractometry.

2 Method ribbon Displays the type of density determination and the method identification or the selected standard.

  • MS : measurement using a method
  • TE : test
  • ADJ : adjustment

3 Sample ID ribbon Displays the sample ID.

4 Measuring ribbon Displays the unit, the measured value or result and the temperature of the measuring cell.

5 Sampling ribbon Displays the current status of the sample pump or syringe.

6 Key functions ribbon Displays the function of the keys.

3.3.2 Icons on the screen

Status ribbon Barcode reader is scanning. Status ribbon RFID reader is reading or writing.

Status ribbon EasyDirect Density & Refractometry is connected.

Status ribbon Density meter is sending data to the printer.

Status ribbon Shows the charge of the battery.

Status ribbon Battery is charging.

Sampling ribbon Syringe mode activated. The sample pump is deactivated.

Sampling ribbon Sample pump activated and piston in the lowermost position.

Sampling ribbon Sample pump activated and piston in the uppermost position.

10 Design and Function Eagle Eye Power Solutions - SG-Pro

Menu Results Start · Icon Location Explanation

Page 13 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 4.2 Unpack the density meter
Page 13 · click to enlarge
Page 13

4 Installation and commissioning

4.1 Scope of delivery

Filling Tube 190 mm

  • 30330847
  • Tube
  • Connector syringe/filling tube
  • Washer
  • 30330855

Power Supply and Worldwide Adapters

  • 30449255
  • USB-C Cable 30449253
  • Density standards (3 pcs) 6 mL 51325005

Tag SmartSample (10 pcs) 30449268 – •

  • User Manual –
  • Declaration of conformity –
  • Test report –

4.2 Unpack the density meter

1 Remove the density meter from the protective packaging. 2 Store the packing material for later transport over long distances. 3 Check if you received all parts listed in the scope of delivery. 4 Inspect the parts visually for flaws or damage. 5 If parts are missing or damaged, report it immediately and file a freight claim if needed.

Show 8 more labels / short text on this page
Part · Order number · SG-Pro · DensitoPro · Handheld density meter · • · • · Battery Li-Ion 2400mAh
Page 14 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 4.3.2 Charge with a computer
Page 14 · click to enlarge
Page 14

4.3 Charge the density meter

You can either charge the density meter using the supplied AC adapter or a computer. Recharging an empty battery takes approximately three hours.

4.3.1 Charge with the AC adapter

The AC adapter is suitable for all supply line voltages ranging from 100...240 V AC and 50/60 Hz.

WARNING

Danger of death or serious injury due to electric shock! Contact with parts that carry a live current can lead to death or injury. 1 Only use the AC adapter designed for your instrument. 2 Keep all electrical cables and connections away from liquids and moisture. 3 Check the cables and the plugs for damage and replace damaged cables and plugs.

NOTICE

Assemble the AC adapter

1 Slide the desired set of prongs (1) into the AC adapter (2). 1

Eagle Microsystems Handheld Density Meter (SG Pro) — NOTICE diagram (page 14)
Fig. 10

2 Insert the USB-A plug of the USB-C cable in the USB-A socket of the AC adapter.

Connect the density meter to the power supply

1 Insert the USB-C plug of the USB-C cable in the USB-C socket of the density meter. 2 Install the cables in such a way that they cannot be damaged or interfere with operation. 3 Insert the plug of the AC adapter in a power outlet that is easily accessible.

The density meter starts charging and the icon is displayed.

Change the prongs of the AC adapter

1 Pull the plug of the AC adapter out of the power outlet. 1 2 Push the button (2) and slide the set of prongs (1) out. 2

3 Slide the desired set of prongs into the AC adapter.

4.3.2 Charge with a computer

  • A running computer with USB-A sockets (USB 2.0 or higher) is available. 1 Insert the USB-C plug of the USB-C cable in the USB-C socket of the density meter. 2 Install the cables in such a way that they cannot be damaged or interfere with operation.
Page 15 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 4.6 Activate and deactivate the barcode reader (DensitoPro only)
Page 15 · click to enlarge
Page 15

3 Insert the USB-A plug in the USB-A socket of the computer.

The density meter starts charging and the icon is displayed.

4.4 Setup for working with the sample pump

4.4.1 Activate the sample pump mode

1 Press the Menu key.

2 Navigate to Settings > Measurement > Sampling.

The Sampling window opens.

3 Navigate to Sample pump.

4 Press the OK key. 5 Navigate back to the home screen. The icon is displayed in the sampling ribbon.

4.4.2 Install the filling tube

  • The filling tube (tube (1), connector (2) and washer (3)) is assembled.
Eagle Microsystems Handheld Density Meter (SG Pro) — 4.4.2 Install the filling tube diagram (page 15)
Fig. 11 · 4.4.2 Install the filling tube
  • The locking screw (2) is installed and tight. −Screw the connector (1) of the filling tube into the
Eagle Microsystems Handheld Density Meter (SG Pro) — 4.4.2 Install the filling tube diagram (page 15)
Fig. 12 · 4.4.2 Install the filling tube

opening (3) and tighten it.

4.5 Activate and deactivate the RFID reader (DensitoPro only)

1 Press the Menu key.

2 Navigate to Settings > RFID / Barcode > Type.

The Type window opens.

3 To activate the RFID reader, navigate to RFID.

4 To deactivate the RFID reader, navigate to None.

See also Read and write information with the RFID reader (DensitoPro only) } Page 44

4.6 Activate and deactivate the barcode reader (DensitoPro only)

1 Press the Menu key.

2 Navigate to Settings > RFID / Barcode > Type.

The Type window opens.

Page 16 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 4.7.1.1 Activate the syringe mode
Page 16 · click to enlarge
Page 16

3 To activate the barcode reader, navigate to Barcode.

4 To deactivate the barcode reader, navigate to None.

See also Read information with the barcode reader (DensitoPro only) } Page 45

4.7 Install accessories

4.7.1 Setup for working with a syringe

When you work with a syringe you can place the density meter either on the bench or hold it upright.

Eagle Microsystems Handheld Density Meter (SG Pro) — 4.7.1 Setup for working with a syringe diagram (page 16)
Fig. 13 · 4.7.1 Setup for working with a syringe
Eagle Microsystems Handheld Density Meter (SG Pro) — 4.7.1 Setup for working with a syringe diagram (page 16)
Fig. 14 · 4.7.1 Setup for working with a syringe

4.7.1.1 Activate the syringe mode

  • The piston is in the lowermost position. 1 Press the Menu key.

2 Navigate to Settings > Measurement > Sampling.

The Sampling window opens.

3 Navigate to Syringe.

4 Press the OK key. 5 Navigate back to the home screen. The icon is displayed in the sampling ribbon. 6 If you want place the density meter on the bench, activate the auto rotation of the screen.

See also Activate and deactivate the auto rotation of the screen } Page 20

Page 17 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 4.7.2 Connect a printer
Page 17 · click to enlarge
Page 17

4.7.1.2 Install the connector for the syringe

  • The connector for the syringe is available.
  • The filling tube is installed. 1 Unscrew the locking screw counterclockwise with a flat screwdriver and remove it.
Eagle Microsystems Handheld Density Meter (SG Pro) — 4.7.1.2 Install the connector for the syringe diagram (page 17)
Fig. 15 · 4.7.1.2 Install the connector for the syringe

2 Screw the connector into the filling opening and tighten it.

Eagle Microsystems Handheld Density Meter (SG Pro) — 4.7.1.2 Install the connector for the syringe diagram (page 17)
Fig. 16 · 4.7.1.2 Install the connector for the syringe

See also Accessories } Page 65 Install the filling tube } Page 13

4.7.2 Connect a printer

In order to print the results right after the measuring procedure, the density meter has to be connected to a USB or Bluetooth printer. Alternatively the results can be printed at a later time from the Results menu.

The following printer types are compatible with the density meter:

Bluetooth Godex MX20 USB P25

The bluetooth function is only available in selected countries with the necessary approvals. For more infor- mation, contact an authorized EEPS dealer or service representative.

Connection type Printer

Page 18 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 4.7.2.3 Print a test page
Page 18 · click to enlarge
Page 18

Refer to the user documentation of your printer for setting up and operating the printer.

See also Accessories } Page 65

4.7.2.1 Connect a USB printer

1 Press the Menu key.

2 Navigate to Settings > Printer > Printer type.

The Printer type window opens.

3 Navigate to USB.

5 Connect the USB-A plug of USB-C cable with the USB-A socket of the printer. 6 Connect the USB-A plug with the USB-A socket of the density meter. 7 Switch on the printer. The connection is established.

See also Configure the automatic printing of results } Page 31 Print results } Page 47

4.7.2.2 Connect a bluetooth printer

  • The bluetooth printer is set up and running. 1 Insert the bluetooth dongle in the USB-A socket of the density meter.

2 Press the Menu key.

Eagle Microsystems Handheld Density Meter (SG Pro) — 4.7.2.2 Connect a bluetooth printer diagram (page 18)
Fig. 17 · 4.7.2.2 Connect a bluetooth printer

3 Navigate to Settings > Printer > Printer type.

The Printer type window opens.

4 Navigate to Bluetooth.

6 Press the Menu key.

7 Navigate to Printer selection.

The Printer selection window with a list of

available bluetooth printers opens. 8 Navigate to the bluetooth printer you want to connect.

Configure the automatic printing of results } Page 31 Print results } Page 47

4.7.2.3 Print a test page

  • The printer is connected and running. 1 Press the Menu key.

2 Navigate to Settings > Printer > Test printer.

If there is no connection yet, the notification Connecting to printer is displayed. When the connection is established, a test page is printed.

See also

Page 19 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 4.8.1 Change the date and time format
Page 19 · click to enlarge
Page 19

See also Connect a printer } Page 15

4.7.3 Connect to EasyDirect Density & Refractometry

EasyDirect Density & Refractometry is a data management software to manage a database of measurement results. By connecting the density meter via a USB-C cable to a computer on which EasyDirect Density & Refractometry is installed, the data can be directly exported to the software.

Operating EasyDirect Density & Refractometry is described in the help of the software.

  • EasyDirect Density & Refractometry is installed on the computer.
  • A USB-C cable is available. 1 Insert the USB-C plug of the USB-C cable in the USB-C socket of the density meter. 2 Insert the USB-A plug in a USB-A socket on the computer. 3 Add the density meter to the list of instruments on EasyDirect Density & Refractometry. The icon is displayed in the status ribbon.

See also Export results to EasyDirect Density & Refractometry } Page 47

4.8 Configure the density meter

4.8.1 Change the date and time format

The following date formats are available:

  • dd/mm/yyyy : day/month/year, e.g. 30/11/2018
  • mm/dd/yyyy : month/day/year, e.g. 11/30/2018
  • yyyy/mm/dd : year/month/day, e.g. 2018/11/30
  • dd.mm.yyyy : day.month.year, e.g. 30.11.2018 The following two time formats are available.
  • 24 h : 24-hour clock, e.g. 6:30 and 18:30
  • am/pm : 12-hour clock, e.g. 6:30 am and 6:30 pm

1 Press the Menu key.

2 Navigate to Settings > Date / Time.

The Date / Time window opens.

3 Navigate to Date or Time.

4 To change the date or time, move the navigation joystick up to raise or down to lower the number. 5 To switch between day, month and year or hours and minutes, move the navigation joystick to the left or right.

7 Navigate to Date format or Time format.

8 Navigate to the format you want to use.

Procedure · Procedure

Page 20 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 4.8.4 Change energy saver settings
Page 20 · click to enlarge
Page 20

4.8.2 Change the temperature unit

Two temperature units are available.

  • °C
  • °F

When you change the temperature unit, temperatures are displayed in the selected unit. The temperature unit of results that were determined before you changed the unit remain unchanged.

1 Press the Menu key.

2 Navigate to Settings > Global units > Temperature unit.

The Temperature unit window opens.

3 Navigate to the desired Temperature unit.

The temperatures are displayed in the selected unit.

4.8.3 Change the density unit

Three density units are available.

  • g/cm³
  • kg/m³
  • lb/gal (US)

When you change the density unit, density values are displayed in the selected unit. There are two exceptions.

  • The densities of results that were determined before you changed the unit remain unchanged.
  • If you select the unit [lb/gal], the unit [g/cm 3 ] is still used for adjustments and tests.

1 Press the Menu key.

2 Navigate to Settings > Global units > Density unit.

The Density unit window opens.

3 Navigate to the desired Density unit.

4 Press the OK key. The densities are displayed in the selected unit

4.8.4 Change energy saver settings

The settings of the parameters Brightness, Energy saver and Auto shutdown influence the power consumption and thus the battery life.

Brightness Defines the brightness of the screen. 10...100 %

Energy saver Defines if the screen is dimmed when no action has been performed for a given time.

Off | 30 | 60 | 90 | 120 seconds

Auto shutdown Defines if the density meter is shut down when no action has

Off | 2 | 5 | 10 minutes

been performed for a given time.

1 Press the Menu key.

2 Navigate to Settings > Screen.

The Screen window opens.

3 Navigate to the desired parameter and change the settings.

Show 4 more labels / short text on this page
Procedure · Procedure · Parameter Description Values · Procedure
Page 21 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 4.8.5.2 Disable the password protection
Page 21 · click to enlarge
Page 21

4.8.5 Protect actions with a password

With the parameter Password you can prevent users from performing certain actions without a password. The following table shows which actions are password-protected.

Start methods. No Start tests. No

Start adjustments. Yes View the configuration of methods, tests and adjustments. No

Change the configuration of methods, tests and adjustments. Yes

Create and delete methods. Yes

View, print and export results. No Delete all results. Yes

View the settings of the density meter. No Change the settings of the density meter. Yes

Update the firmware. Yes Restore factory adjustment. Yes

Restore factory settings. Yes

The password has to be entered once before a user performs a password-protected action and does not have to be reentered until the density meter is restarted. If the user forgets the password, the universal password 8606 unlocks all actions and settings.

4.8.5.1 Enable the password protection

1 Press the Menu key.

2 Navigate to Settings > Password > Password protection.

The Password protection window opens.

3 Navigate to Yes.

The Password window opens.

5 Enter a password with the navigation joystick.

6 Press the OK key. 7 Reenter the password.

The password has to be entered before the user can perform a protected action.

4.8.5.2 Disable the password protection

1 Press the Menu key.

2 Navigate to Settings > Password > Password protection.

The Password protection window opens.

3 Navigate to No.

The Password window opens. 5 Enter the password with the navigation joystick.

6 Press the OK key. Users can perform all actions.

Action Password protected

Page 22 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 4.8.10 Activate and deactivate a cleaning reminder
Page 22 · click to enlarge
Page 22

4.8.5.3 Change the password

1 Disable the password protection. 2 Enable the password protection and enter the new password.

4.8.6 Activate and deactivate the auto rotation of the screen

The Auto rotate parameter defines whether the screen rotates by 180° when the density meter is placed on a bench. 1 Press the Menu key.

2 Navigate to Settings > Screen > Auto rotate.

The Auto rotate window opens.

3 Navigate to Yes to switch on or No to switch off the Auto rotate mode.

4 Press the OK key. The screen rotates when the density meter is placed on a bench.

4.8.7 Activate and deactivate the audio signal

The parameter Audio signal defines whether an audio signal is played when a key or the navigation joystick is pressed and when measurements are finished. 1 Press the Menu key.

2 Navigate to Settings > Audio signal.

The Audio signal window opens.

3 Navigate to Yes to switch on or No to switch off the Audio signal.

4.8.8 Change the keyboard layout

1 Press the Menu key.

2 Navigate to Settings > Language / Keyboard > Keyboard layout.

The Keyboard layout window opens.

3 Navigate to the desired keyboard layout.

4.8.9 Change the language

The system language defines the language of the user interface and printouts.

1 Press the Menu key.

2 Navigate to Settings > Language / Keyboard > Language.

The Language window opens. 3 Navigate to the desired language.

4 Press the OK key. The user interface is changed to the selected language.

4.8.10 Activate and deactivate a cleaning reminder

Because the measuring cell always contains residue from the last sample, there is a risk that the measuring cell can be damaged by the sample. To prevent such damages, EEPS recommends that you clean the measuring cell at the end of each sample series.

To remind users to clean the measuring cell, you can activate a reminder. This reminder is displayed before manual shutdown of the instrument, and users have to confirm or cancel the reminder before shutting down the density meter.

Page 23 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 4.9.1 Create, change or delete a method
Page 23 · click to enlarge
Page 23

Action of the system Cleaning reminder Auto shutdown

When users shut down the density meter, a reminder is displayed.

If the density meter automatically shuts down, no reminder is displayed.

No reminder is displayed when the density meter shuts down.

No Yes/No

1 Press the Menu key.

2 Navigate to Settings > Measurement > Cleaning reminder.

The Cleaning reminder window opens.

3 Navigate to Yes to activate or No to deactivate the Cleaning reminder.

See also Change energy saver settings } Page 18

4.9 Configure density determinations

Three different types of density determinations are available.

  • Methods : determine the density of a sample.
  • Test : determine the density of a standard to check that the measurement cell is clean and measuring accurately.
  • Adjustment : determine the density of a standard and adjust the density meter based on the results.

4.9.1 Create, change or delete a method

4.9.1.1 Create a method

  • The new method is listed in the Methods window with the name you enter in the parameter Method ID . If you do not change the parameter Method ID , it is set by default to (A{X}{X}), e.g. A03.
  • Less than 30 methods are saved.
  • The home screen is open. 1 Move the navigation joystick to the left.

The Methods window opens.

2 Press the New key.

The Method settings window opens.

3 Navigate to Method ID and enter a name with up to 25 characters.

The method is listed in the Methods window. 5 Change the parameters as needed.

See also Configure the user guidance } Page 23 Configure the measurement reliability } Page 24 Configure the result conversion } Page 24 Configure the automatic documentation of results } Page 30 Configure result limits } Page 32

Yes No · Yes Yes · Procedure

Page 24 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 4.9.3 Change an adjustment
Page 24 · click to enlarge
Page 24

4.9.1.2 Change a method

  • The home screen is open. 1 Move the navigation joystick to the left.

The Methods window opens.

2 Navigate to the method you want to edit.

The Method settings window opens. 3 Change the parameters as needed.

See also Configure the user guidance } Page 23 Configure the measurement reliability } Page 24 Configure the result conversion } Page 24 Configure the automatic documentation of results } Page 30 Configure result limits } Page 32

4.9.1.3 Delete a method

  • More than one method is created.
  • The method you want to delete is not active, marked by a square.
  • The home screen is open. 1 Move the navigation joystick to the left.

The Methods window opens. 2 Navigate to the method you want to delete.

The Method settings window opens.

3 Press the Delete key.

The message Do you want to delete the method? is displayed.

4 Navigate to Yes.

5 Press the OK key. The method is no longer listed in the Methods window.

4.9.2 Change a test

1 Press the Menu key.

2 Navigate to Test.

The Test window opens.

3 Change the parameters as needed.

See also Configure the user guidance } Page 23 Configure the automatic documentation of results } Page 30 Configure a standard } Page 32 Configure the tolerance for a test } Page 33

4.9.3 Change an adjustment

1 Press the Menu key.

2 Navigate to Adjustment.

The Adjustment window opens.

3 Change the parameters as needed.

Page 25 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 4.9.4.4 Define the cleaning type (only guided workflow)
Page 25 · click to enlarge
Page 25

See also Configure the user guidance } Page 23 Configure the automatic documentation of results } Page 30 Configure a standard } Page 32

4.9.4 Configure the user guidance

The parameters that are available to configure the user guidance are described in the following chapters.

4.9.4.1 Define the level of user guidance

The Workflow parameter defines the level of user guidance during the density determination. Two options are available.

  • Manual : users decide how and when to fill the measuring cell, start the measurement and drain or clean the measuring cell.
  • Guided : the density meter prompts the user to fill, drain or clean the measuring cell.

4.9.4.2 Define the pump speed

The parameter Fill speed defines the speed of the pump for filling the measuring cell. The sample viscosity places limits on the fill speed.

Speed level % of maximum pump speed

Fill time for water Viscosity

High 80 % Approx. 4 s Up to 40 mPa*s

Medium 40 % Approx. 8 s Up to 75 mPa*s

Low 10 % Approx. 30 s Up to 100 mPa*s

4.9.4.3 Define the fill cycle (only guided workflow)

The parameter Fill cycle defines, how many times the measuring cell is filled and emptied before the density is measured. The sample that is in the measuring cell after the last filling is used for the measurement. Recommendation:

  • Density determination with the sample pump: set Fill cycle to 3.
  • Density determination with a syringe: set Fill cycle to 2.

4.9.4.4 Define the cleaning type (only guided workflow)

Two types of cleaning methods are available, Drain and Rinse.

Drain The user is prompted to drain the measuring cell after the measurement. This setting is useful if the samples that are measured dissolve in each other.

The user is prompted to fill and drain the measuring cell with a cleaning solution after the measurement. The parameter Rinse cycle defines how many times the user is prompted to fiil and drain the measuring cell. This setting is only useful if the samples that are measured do not dissolve in each other.

See also Typical phases of density determinations } Page 36

Rinse

Page 26 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 4.9.6 Configure the result conversion
Page 26 · click to enlarge
Page 26

4.9.4.5 Activate and deactivate the confirmation of results

After the measurement reliability is achieved, the results are frozen. The parameter Confirmation of results defines whether users have to confirm the results before they are replaced by the current density value. Regardless of the setting, all results are stored. The setting is not method specific but valid for all density determinations.

Action of the system Confirmation of results Workflow

The screen is frozen and the sample pump is locked until users confirm the result.

The screen is frozen until users confirm the result but the sample pump is not locked.

Users are prompted to place the filling tube into the waste beaker. As soon as users press OK, the result is replaced by the current density value.

As soon as the user starts draining the cell, the result is replaced by the current density value.

1 Press the Menu key.

2 Navigate to Settings > Measurement > Confirmation of results.

The Confirmation of results window opens.

3 Navigate to Yes to activate or No to deactivate the Confirmation of results.

See also Define the level of user guidance } Page 23

4.9.5 Configure the measurement reliability

Measurement reliability is only available for methods.

The parameter Measurement reliability defines how the stability of a measured value is evaluated. Once a measured value is determined to be stable, it is saved as a result.

  • Medium : The result is saved as soon as the measured temperature value stays within 0.4 °C (0.72 °F) for 10 s. Results are delivered faster, but with less accuracy compared to the High mode.
  • High : The result is saved as soon as the measured temperature value stays within 0.2 °C (0.36 °F) for 10 s. Results are very accurate, but may take a longer time compared to the Medium mode in case s where the sample temperature differs from the ambient temperature.
  • Manual (only for unguided density determinations): users decide when the result is stable and save it.

Define the level of user guidance } Page 23

4.9.6 Configure the result conversion

The density meter converts the measured oscillation to the unit you define. The conversions are grouped in the following categories.

  • Density : Density and specific gravity
  • Alcohol : Alcohol concentration
  • API : American Petroleum Institute (API density, API specific gravity, API degree)
  • Baumé : Baumé scale
  • H2SO4 : Sulfuric acid concentration
  • Sugar : Sucrose concentration
Show 6 more labels / short text on this page
Yes Guided · Yes Manual · No Guided · No Manual · Procedure · See also
Page 27 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 8,(Y/Y)" I (8FRUQJ) I ( - 2 4 )
Page 27 · click to enlarge
Page 27
  • User-defined : Conversion as defined by the user

Select the category and the conversion type

1 Press the Menu key.

2 Navigate to Methods > Method settings > Calculation > Category

3 Navigate to the desired category.

The Calculation window opens. 5 Navigate to the conversion type you want to use.

6 Press the OK key. The parameters that you need to configure for the selected conversion type are displayed. You can find more information about the conversion types and the parameters in the following chapters.

4.9.6.1 Density and specific gravity

Available conversions:

  • d : density at the measurement temperature
  • d comp. : temperature compensated density
  • SG : temperature compensated specific gravity
  • SG (t/t) : specific gravity at the measurement temperature

Density at the measurement temperature

Unit: [g/cm3], [kg/m3] or [lb/gal (US)] as defined in Settings > Global units. The calculation is based on a reference for the density of water.

Unit: [g/cm3], [kg/m3] or [lb/gal (US)] as defined in Settings > Global units.

The measured density is converted to the value it would have at a compensation temperature. The following formula is used to compensate the temperature influence. IHTRU"IRJFX*(1+ȼ*(9RJFX-9HTRU ))

  • d comp : temperature compensated density
  • d meas : density a the measurement temperature
  • α : temperature coefficient as defined in Alpha*1000 > Alpha entry , see [Define the temperaturecompensation coefficient α } Page 29]
  • T meas : measurement temperature, temperature of the measuring cell during the measurement
  • T comp : compensation temperature as defined in Compensation temp.

Temperature compensated specific gravity

8," IHTRU(8FRUQJ) IHTRU(-24)

  • SG: specific gravity
  • d comp (Sample): density of the sample compensated to the temperature defined in Compensation temp. 1 . The temperature influence is compensated by the temperature coefficient α as defined in Alpha*1000 > Alpha entry . See [Define the temperature-compensation coefficient α } Page 29]
  • d comp (H 2 O): density of water compensated to the temperature defined in Compensation temp. 2

Specific gravity at the measurement temperature

8,(Y/Y)" I (8FRUQJ) I ( - 2 4 )

Temperature compensated density

Page 28 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 4.9.6.4 Baumé scale
Page 28 · click to enlarge
Page 28
  • SG(t/t): specific gravity
  • d(Sample): density of the sample at the measurement temperature.
  • d(H 2 O): density of water compensated to the measurement temperature.

See also Define the temperature-compensation coefficient α } Page 29 Configure the result conversion } Page 24 Density of pure water (0…40 °C) } Page 68 Change the density unit } Page 18

4.9.6.2 Alcohol concentration

Available calculations:

  • Alcohol [%w/w] : mass percent [%w/w]
  • Alcohol 60°F [%v/v] : volume percent [%v/v], compensated to 60 °F
  • Alcohol 20°C [%v/v] : volume percent [%v/v], compensated to 20 °C
  • Proof US 60°F [Proof] : Proof degree with US unit [Proof], compensated to 60 °F
  • Proof IP 60°F [Proof] : Proof degree with IP unit [Proof], compensated to 60 °F

See also Configure the result conversion } Page 24

4.9.6.3 American Petroleum Institute (API)

The conversion of the measured value to the selected unit is based on the tables published by the American Petroleum institute (API). The following tables are available.

  • Crude oil (A)
  • Refined products (B)
  • Lubricants (D)

The Compensation temp. parameter defines the compensation temperature.

The Output format defines to which of the following units the result is converted:

  • d : API density
  • SG : API specific gravity
  • °API : API degree

See also Configure the result conversion } Page 24

4.9.6.4 Baumé scale

Unit[°Bé] Available conversions:
  • H. Bé 15°C [°Bé] : for densities > 1 g/cm 3 , compensated to 15 °C
  • L. Bé 15°C [°Bé] : for densities < 1 g/cm 3 , compensated to 15 °C
  • H. Bé 60°F [°Bé] : for densities > 1 g/cm 3 , compensated to 60 °F
  • L. Bé 60°F [°Bé] : for densities < 1 g/cm 3 , compensated to 60 °F

The temperature influence is compensated by the temperature coefficient α as defined in Alpha*1000 > Alpha entry.

Page 29 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 4.9.6.7 User defined concentration
Page 29 · click to enlarge
Page 29

See also Define the temperature-compensation coefficient α } Page 29 Configure the result conversion } Page 24

4.9.6.5 Sulfuric acid concentration

Available conversions:

  • H2SO4 [%w/w] : mass percent [%w/w]
  • H2SO4 20°C [%v/v] : volume percent [%v/v], compensated to 20 °C

See also Configure the result conversion } Page 24

4.9.6.6 Sugar concentration

Available conversions:

  • Brix [%w/w] : mass percent [w/w %], compensated to 20 °C
  • Plato [°P] : degree Plato [°P], compensated to 20 °C

See also Configure the result conversion } Page 24

4.9.6.7 User defined concentration

You can define a function to convert a density or specific gravity to a concentration. The following steps are involved:

1. Define the conversion of the measured value to a density or specific gravity.
2. Define the function to convert the density or specific gravity to a concentration. 3. Define the unit and the number of decimal places of the concentration.

Define the density or specific gravity } Page 27 Define the function } Page 28 Define the format of the result } Page 29 Configure the result conversion } Page 24

4.9.6.7.1 Define the density or specific gravity

The conversions are either based on temperature compensated densities (d comp.) or specific gravities (SG).

Define a temperature-compensated density

The following formula is used to calculate the temperature-compensated density. IHTRU"IRJFX*(1+ȼ*(9RJFX-9HTRU ))

  • d comp : temperature compensated density
  • d meas : density a the measurement temperature
  • α : temperature coefficient as defined in Alpha*1000 > Alpha entry , see [Define the temperaturecompensation coefficient α } Page 29]
  • T meas : measurement temperature, temperature of the measuring cell during the measurement
  • T comp : compensation temperature as defined in Compensation temp. Steps to define a temperature-compensated density:
  • The conversion category is set to User-defined . 1 Navigate to X and select one of the following options: d comp. , d comp. - 1 or (1/d comp.) - 1 .

See also

Page 30 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 4.9.6.5 Sulfuric acid concentration
Page 30 · click to enlarge
Page 30

2 Navigate to Compensation temp. and define the compensation temperature.

3 Navigate to Alpha*1000 > Alpha entry and define α. See [Define the temperature-compensation coefficient α } Page 29]

Define a temperature-compensated specific gravity The following formula is used to calculate the temperature-compensated specific gravity.

8," IHTRU(8FRUQJ) IHTRU(-24)

  • SG: specific gravity
  • d comp (Sample): density of the sample compensated to the temperature defined in Compensation temp. 1. The temperature influence is compensated by the temperature coefficient α as defined in Alpha*1000 > Alpha entry . See [Define the temperature-compensation coefficient α } Page 29]
  • d comp (H 2 O): density of water compensated to the temperature defined in Compensation temp. 2 Steps to define the temperature-compensated specific gravity:
  • The conversion category is set to User-defined . 1 Navigate to X and select one of the following options: SG , SG-1 or (1/SG)-1 .

2 Navigate to Compensation temp. 1 and define the compensation temperature for the sample.

3 Navigate to Compensation temp. 2 and define the compensation temperature for water.

4 Navigate to Alpha*1000 > Alpha entry and define α. See [Define the temperature-compensation coefficient α } Page 29]

See also Define the function } Page 28

4.9.6.7.2 Define the function

You can either use a linear function (Linear (Ax+B)) or a second order polynomial function (2nd order pol. (ax²+bx+c). The functions are based on a graph with the density or specific gravity on the x-axis and the concentration on the y-axis. The following formulae are used.

  • Linear function: Ax + B
  • Second order polynomial function: Ax 2 + Bx + C
  • Procedure
  • The coefficients A and B are known for a linear function.
  • The coefficients A, B and C are known for a second order polynomial function. 1 Navigate to Fit type and select Linear (Ax+B) or 2nd order pol. (ax²+bx+c) .

3 Enter the values for the coefficients.

Calculate the coefficients A, B and C 1 If you want to use a second order polynomial function, prepare at least 4 solutions with different concentrations. 2 If you want to use a linear function, prepare at lest 2 solutions with different concentrations. 3 Determine the density or specific gravity you want to use for each concentration. 4 Plot the concentration (y-axis) against the density or specific gravity (x-axis). 5 Fit a line or second order polynomial to the curve.

Define the density or specific gravity } Page 27

See also

Page 31 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — ȼ"{ I 2 -1) / {9 -9 ) I 1
Page 31 · click to enlarge
Page 31

4.9.6.7.3 Define the format of the result

Using the parameters Unit and Decimal places, you can define the unit and the number of decimal places for the display of the current value and the result.

4.9.6.8 Define the temperature-compensation coefficient α

The temperature-compensation coefficient α is used to convert a measured value to a reference temperature. You have three options to define α.

  • Manual : users enter the value for α that they calculated or read from a table.
  • Calculated : the density meter calculates α based on a linear equation and densities and temperatures that users enter.
  • 2nd order pol. (ax²+bx+c) : the density meter calculates α based on a second order polynomial using coefficients that users enter.

See also Density and specific gravity } Page 25 Baumé scale } Page 26 Define the density or specific gravity } Page 27

4.9.6.8.1 Enter a known value for α

  • The result of α *1000 is known.

1 Navigate to Alpha*1000 > Alpha entry and select Manual.

3 Navigate to Alpha*1000. 4 Enter the value for α*1000.

1 Measure the density (d1) at a temperature (T1) above the normal measuring temperature. 2 Measure the density (d2) at a temperature (T2) below the normal measuring temperature. 3 Use the measured densities and the temperatures in [°C] to calculate α according to the following formula. 4 Calculate the result of α*1000.

ȼ"{ I 2 -1) / {9 -9 ) I 1

Example Density at 26 °C (T1) = 0.7844 g/cm3

Density at 15 °C (T2) = 0.7937 g/cm3

See also Temperature-compensation coefficients α } Page 68

4.9.6.8.2 The density meter calculates a fixed value for α

  • The density (d 1 ) at a temperature (T 1 ) above the normal measuring temperature is known.
  • The density (d 2 ) at a temperature (T 2 ) below the normal measuring temperature is known. 1 Navigate to Alpha*1000 > Alpha entry and select Calculated .

2 Navigate to Density 1 and enter the value for d1. 3 Navigate to Temperature 1 and enter the value for T1.

Calculate α

Page 32 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 4.9.7.1 Configure the sample or standard identification
Page 32 · click to enlarge
Page 32

4 Navigate to Density 2 and enter the value for d2. 5 Navigate to Temperature 2 and enter the value for T2. The density meter calculates the value for Alpha*1000 and displays it.

See also List of errors and problems } Page 57

4.9.6.8.3 Define the second order polynomial used to calculate α

The density meter can use the following second order polynomial to calculate α.

2 +G*I RJFX +H)

Procedure ▪ The three coefficients a, b and c are known. 1 Navigate to Alpha*1000 > Alpha entry and select 2nd order pol. (ax²+bx+c).

3 Enter the values for the coefficients a, b and c.

Calculate the coefficients a, b and c 1 Prepare at least 4 solutions with different concentrations. 2 Define two temperatures, one above und one below the normal measurement temperature. 3 Measure the density of each solution at both temperatures. 4 Calculate the temperature-compensation coefficient α for each solution. 5 Plot the values for α (y-axis) against the densities (x-axis) measured at one of the temperatures. 6 Fit a second order polynomial to the curve.

See also List of errors and problems } Page 57

4.9.7 Configure the automatic documentation of results

NOTICE

Loss of data! A maximum of 1100 results is saved in the density meter. When the result list contains 1100 results and a new results is saved, the oldest result is overwritten. − Print or export the results before the maximum number of results is reached.

See also Change a method } Page 22 Change a test } Page 22 Change an adjustment } Page 22

4.9.7.1 Configure the sample or standard identification

You can define what kind of identification is assigned to a sample or to a standard.

Automatic numbering is only available for methods.

Sample ID Standard ID Sample ID entry Standard ID entry Action of the system

No identification is assigned to the sample or the standard.

ȼ"{F*IRJFX · None Not available

Page 33 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 4.9.7.2 Configure the automatic printing of results
Page 33 · click to enlarge
Page 33

Sample ID Standard ID Sample ID entry Standard ID entry Action of the system

  • The identification you define in the Sample ID parameter is assigned to all samples measured with the same method.
  • The identification you define in the Standard ID parameter is assigned to all standards measured with the same test.

Users enter the identification before the density measurement starts.

  • Guided density determination: users are prompted to enter the identification when the method starts. The density determination continues when the user has entered a valid identification.
  • Unguided density determination: Users have to enter the identification before the measurement starts but are not prompted to do so. If they do not enter an identification, the identification of the previous sample is used.

An automatically generated identification is assigned to the sample. The identification automatically increases by 1, when a new measurement is started with the same method.

Automatic numbering Not available

4.9.7.2 Configure the automatic printing of results

The Printout parameter defines whether a result is automatically printed when it is saved and which infor- mation is printed.

  • None : No report is printed.
  • Short : A short report is printed.
  • Long : A long report is printed

Information Methods Test Adjustment

Date / Time Short/Long Short/Long Short/Long

Method ID Short/Long – –

Sample ID Short/Long – –

Application name with standard

– Short/Long Short/Long

Standard ID – Short/Long Short/Long

Status Short/Long Short/Long –

Calculation Short/Long – –

Density Short/Long Short/Long Short/Long

Dev. d – Short/Long Short/Long

Temp. Short/Long Short/Long Short/Long

d nominal – Short/Long Short/Long

SNR Long Long Long

Fixed 0…25 characters · Variable Not available

Page 34 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 4.9.9 Configure a standard
Page 34 · click to enlarge
Page 34

Information Methods Test Adjustment

Additional settings like the compensation temperature and the temperature-compen-

See also Connect a printer } Page 15

4.9.7.3 Configure the automatic export of results

Results are automatically exported to EasyDirect Density & Refractometry if the density meter is connected to a computer with the software installed.

See also Connect to EasyDirect Density & Refractometry } Page 17

4.9.8 Configure result limits

Result limits are only available for methods.

The Result limits parameter defines whether the density meter evaluates if a result lies within a defined range. The parameters Lower limit and Upper limit define the range.

  • Result within the limits: the background of the measuring ribbon turns green.
  • Result outside the limits: the background of the measuring ribbon turns red and the text Out of limits is displayed in the sampling ribbon

1 Press the Menu key.

2 Navigate to Methods > Method settings > Calculation > Result limits.

3 To deactivate Result limits, navigate to No.

4 To activate Result limits, navigate to Yes.

Lower limit and Upper limit are displayed.

6 Set the Lower limit and Upper limit.

4.9.9 Configure a standard

Use water as standard ▪ The Test window or the Adjustment window are open. 1 Navigate to Standard type and select Water.

The density meter uses predefined values for the Test or the Adjustment.

Use another substance as standard ▪ The Test window or the Adjustment window are open. 1 Navigate to Standard type and select Other standard.

3 Navigate to Compensation temp. and enter the compensation temperature.

4 Navigate to d nominal and enter the nominal density of the standard at the temperature you entered as compensation temperature. 5 Navigate to Alpha*1000 and define the temperature-compensation coefficient α.

Long Long Long · sation coefficient α · Procedure

Page 35 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 4.9.10 Configure the tolerance for a test
Page 35 · click to enlarge
Page 35

See also Define the temperature-compensation coefficient α } Page 29

4.9.10 Configure the tolerance for a test

The Tolerance d parameter is used to evaluate the measured density and decide if the test passed or failed.

Recommendations for defining the value for Tolerance d:

  • Deionized water: 2*(instrument resolution + operator repeatability)
  • Certified organic standards with a high temperature-compensation coefficient: uncertainty of the standard + limit of error of instrument + temperature error + 2*operator repeatability

The operator repeatability is the standard deviation of ten density determinations performed by the operator.

Page 36 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 5.2 Shut down the density meter
Page 36 · click to enlarge
Page 36

5 Operation

5.1 Start up the density meter

− Press key 1 (1).

Eagle Microsystems Handheld Density Meter (SG Pro) — 5 Operation diagram (page 36)
Fig. 18 · 5 Operation

A short beep is played. A welcome message is displayed.

The home screen opens. The density meter is ready

to use.

5.2 Shut down the density meter

− Press and hold any of the keys or the navigation joystick for more than 3 s.

A beep is played and the screen goes dark.

The control circuits for the keys and the navigation joystick are energized. The rest of the density meter is

no longer energized.

Shut down using the Menu ▪ The home screen is open. 1 Press the Menu key.

2 Navigate to Shut down.

The dialog Do you want to shut down the instrument? opens.

3 Press the Yes key. A beep is played and the screen goes dark.

The control circuits for the keys and the navigation joystick are energized. The rest of the density meter is

no longer energized.

See also Activate and deactivate a cleaning reminder } Page 20

34 Operation Eagle Eye Power Solutions - SG-Pro

Page 37 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 5.3.2 Enter text and numbers
Page 37 · click to enlarge
Page 37

5.3 Navigate and enter information

5.3.1 Navigate through menus and confirm settings

1 Press the Menu key. 2 To navigate to a desired menu, move the navigation joystick up or down. 3 To navigate to a submenu of the selected menu, move the navigation joystick to the right. 4 To go back, move the navigation joystick to the left.

Eagle Microsystems Handheld Density Meter (SG Pro) — 5.3.1 Navigate through menus and confirm settings diagram (page 37)
Fig. 19 · 5.3.1 Navigate through menus and confirm settings

In every menu, where the function of key 3 is marked with a green bar, you can press the navigation joystick to confirm the selection.

Shortcut to the method list

  • The home screen is open. − Move the navigation joystick to the left.

The Methods window opens.

5.3.2 Enter text and numbers

  • The keyboard window is open. 1 Navigate to the character, number, symbol or icon. 2 Press the navigation joystick to confirm the selection of the character, number, symbol or icon.

3 To confirm the entered information and leave the keyboard window, press the OK key.

Eagle Microsystems Handheld Density Meter (SG Pro) — 5.3.2 Enter text and numbers diagram (page 37)
Fig. 20 · 5.3.2 Enter text and numbers

Min. 1 Max. 25

1 5 2 3

No. Icon Function

Cycles between the screens with small, capital or special characters.

2 Deletes all entered characters.

3 Deletes the character to the left of the cursor.

Operation 35 Eagle Eye Power Solutions - SG-Pro

Show 4 more labels / short text on this page
Confirm settings · Keyboard overview · Method ID · Cancel OK
Page 38 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 5.5 Example: Density determination using the sample pump
Page 38 · click to enlarge
Page 38

No. Icon Function

4 Moves the cursor.

5. Displays available characters, numbers and symbols.
6. Maximum number of characters you can enter.
7. Minimum number of characters you need to enter.
8. Displays the entered characters.

See also Change the keyboard layout } Page 20

5.4 Typical phases of density determinations

Density determinations include the following two phases.

  • Fill the measuring cell with sample and perform the measurement.
  • Prepare the measuring cell for the next sample.

5.4.1 Fill the measuring cell and measure the density

Because the measuring cell always contains residue from the previous sample or the cleaning solution, it is important that you remove the residue using the new sample. This is only possible if the residue is soluble in the new sample.

To achieve good results, make sure that the measuring cell contains only the sample you want to measure and that there are no bubbles in the measuring cell.

You can use the sample pump or syringes to fill, drain and rinse the measuring cell. The sample pump is used for samples with a viscosity up to 100 mPa*s. Syringes are typically used if the viscosity of the samples is above 100 mPa*s.

See also Example: Density determination using the sample pump } Page 36 Example: Density determination using a syringe } Page 39 Example: Density determination of cold or hot samples } Page 42

5.4.2 Prepare the measuring cell for the next sample

At the end of this phase, the residue in the measuring cell must be soluble in the next sample. The solubility of the sample defines if you can drain the measuring cell or if you need to rinse it.

Drain If the next sample dissolves the sample you just measured, you can drain the cell.

If the next sample does not dissolve the sample you just measured, you have to rinse the measuring cell with a cleaning solution. The cleaning solution must dissolve the sample that you just measured and the next sample. When you rinse the measuring cell, the cleaning solution removes the residue from the original sample.

See also Clean the measuring cell } Page 50

5.5 Example: Density determination using the sample pump

The following chapters show you how to configure and perform a density determination of deionized water at room temperature (23 °C or 73.4 °F).

36 Operation Eagle Eye Power Solutions - SG-Pro

Rinse

Page 39 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 5.5.1 Configure the method
Page 39 · click to enlarge
Page 39

Summary of the configuration

  • Guided density determination
  • Filling and draining with the sample pump
  • Evaluation if result lies within specified limits

You can find more information about the configuration of a method in the chapter [Configure density deter- minations } Page 21].

5.5.1 Configure the method

  • The density unit is set to [g/cm 3 ].
  • Confirmation of results in Settings is activated.
  • The home screen is open. 1 Move the navigation joystick to the left.

The Methods window opens.

2 Press the New key.

The Method settings window opens.

3 Navigate to Method ID and enter a name with up to 25 characters.

The method is listed in the Methods window. 5 Change the parameter settings to the values shown in the following table.

Workflow Guided The density meter prompts users to fill and drain the measuring cell.

Sample ID entry Variable The density meter prompts users to enter the sample ID before the density is measured.

Fill cycle 3 The density meter prompts users to fill and drain the measuring cell twice and then fill the measuring cell for the measurement.

Fill speed High The pump fills the measuring cell with 80 % of the maximum pump speed.

Measurement reliability Medium The result is saved as soon as the measured temperature value stays within 0.4 °C (0.72 °F) for 10 s.

Calculation > Category Density A density is calculated.

d The density is not compensated to a certain temperature.

Calculation > Result limits

Yes The density meter evaluates if the result is within a defined range.

Lower limit 0.9972 g/cm3 Defines the lower limit of the range.

Upper limit 0.9977 g/cm3 Defines the upper limit of the range.

Clean Drain Users are prompted to drain the measuring cell.

Printout None The result is not printed.

See also Change the density unit } Page 18 Activate and deactivate the confirmation of results } Page 24 Configure density determinations } Page 21

Operation 37 Eagle Eye Power Solutions - SG-Pro

Parameter Setting Explanation · Calculation > Calcu- lation

Page 40 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 5.5.2 Perform the density determination
Page 40 · click to enlarge
Page 40

5.5.2 Perform the density determination

  • The density meter is set up to work with the sample pump.
  • The piston is in the lowermost position.
  • The home screen is open. 1 Move the navigation joystick to the left. 2 Navigate to the method you want to use and press the Start key. The home screen opens.

3 Press the Start key.

The Sample ID window opens. 4 Enter the identification.

Rinse the measuring cell

1 Immerse the filling tube into the sample (1).

Eagle Microsystems Handheld Density Meter (SG Pro) — 5.5.2 Perform the density determination diagram (page 40)
Fig. 21 · 5.5.2 Perform the density determination

The sample pump starts and stops when the

piston is in the uppermost position. 3 Hold the filling tube over the waste beaker (2).

Eagle Microsystems Handheld Density Meter (SG Pro) — 5.5.2 Perform the density determination diagram (page 40)
Fig. 22 · 5.5.2 Perform the density determination

The sample pump starts and stops when the

piston is in the lowermost position. 5 Clean the end of the filling tube with a clean tissue. 6 Repeat the steps once.

Fill the measuring cell and measure the density

1 Immerse the filling tube into the sample (1).

The sample pump starts and stops when the

piston is in the uppermost position. 3 Make sure there are no bubbles in the measuring cell. The measurement starts.

Eagle Microsystems Handheld Density Meter (SG Pro) — 5.5.2 Perform the density determination diagram (page 40)
Fig. 23 · 5.5.2 Perform the density determination

When the background of the measuring ribbon

changes color, the measurement is finished. The measured value is saved in the results.

38 Operation Eagle Eye Power Solutions - SG-Pro

Start the method

Page 41 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 5.6.1 Configure the method
Page 41 · click to enlarge
Page 41

Drain the measuring cell

1 Hold the filling tube over the waste beaker (1).

Eagle Microsystems Handheld Density Meter (SG Pro) — 4.9.7.3 Configure the automatic export of results diagram (page 41)
Fig. 24 · 4.9.7.3 Configure the automatic export of results

The sample pump starts and stops when the

piston is in the lowermost position. 3 Clean the end of the filling tube with a clean tissue.

5.6 Example: Density determination using a syringe

The following chapters show you how to configure and perform a density determination of a viscous sample such as 40 % w/w sugar solution at room temperature (23 °C or 73.4 °F).

Summary of the configuration

  • Guided density determination
  • Filling and draining with a syringe

You can find more information about the configuration of a method in the chapter [Configure density deter- minations } Page 21].

5.6.1 Configure the method

  • Confirmation of results in Settings is activated.
  • The home screen is open. 1 Move the navigation joystick to the left.

The Methods window opens.

2 Press the New key.

The Method settings window opens.

3 Navigate to Method ID and enter a name with up to 25 characters.

The method is listed in the Methods window. 5 Change the parameter settings to the values shown in the following table.

Workflow Guided The density meter prompts users to fill and drain the measuring cell.

Sample ID entry Variable The density meter prompts users to enter the sample ID before the density is measured.

Fill cycle 2 The density meter prompts users to rinse the measuring cell and then fill the measuring cell for the measurement.

Fill speed High Has no effect because the syringe mode is activated.

Measurement reliability Medium The result is saved as soon as the measured temperature value stays within 0.4 °C (0.72 °F) for 10 s.

Operation 39 Eagle Eye Power Solutions - SG-Pro

Parameter Setting Explanation

Page 42 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 5.6.2 Perform the density determination
Page 42 · click to enlarge
Page 42

Calculation > Category Sugar A sugar concentration is calculated.

Brix [%w/w] A Brix value is calculated.

Calculation > Result limits

None The density meter does not evaluate if the result is within a defined range.

Clean Drain Users are prompted to drain the measuring cell.

Printout None The result is not printed.

Activate and deactivate the confirmation of results } Page 24

5.6.2 Perform the density determination

  • The density meter is set up to work with a syringe.
  • The measuring cell is drained.
  • The home screen is open. 1 Move the navigation joystick to the left.

2 Navigate to the method you want to use and press the Start key. The home screen opens.

3 Press the Start key.

The Sample ID window opens. 4 Enter the identification.

Rinse the measuring cell

1 Fill a new syringe with sample. 2 Insert the syringe into the connector for the syringe.

Eagle Microsystems Handheld Density Meter (SG Pro) — 5.6.2 Perform the density determination diagram (page 42)
Fig. 25 · 5.6.2 Perform the density determination

40 Operation Eagle Eye Power Solutions - SG-Pro

Show 4 more labels / short text on this page
Parameter Setting Explanation · Calculation > Calcu- lation · See also · Start the method
Page 43 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 4.9.7.3 Configure the automatic export of results
Page 43 · click to enlarge
Page 43

3 Hold the filling tube over the waste beaker (1). 4 Press the syringe plunger continuously into the syringe.

Eagle Microsystems Handheld Density Meter (SG Pro) — 4.9.7.3 Configure the automatic export of results diagram (page 43)
Fig. 26 · 4.9.7.3 Configure the automatic export of results

The sample flows through the measuring cell

and removes the residue from the previous sample. 5 Pull the plunger continuously out of the syringe. The sample is pulled back into the syringe and

the measuring cell fills with air. 6 Remove the syringe and empty it into a suitable waste beaker. 1

Fill the measuring cell and measure the density

1 Fill a new syringe with sample. 2 Insert the syringe into the connector for the syringe.

Eagle Microsystems Handheld Density Meter (SG Pro) — 4.9.7.3 Configure the automatic export of results diagram (page 43)
Fig. 27 · 4.9.7.3 Configure the automatic export of results

3 Hold the filling tube over the waste beaker (1). 4 Press the syringe plunger slowly (5…10 cm/s) and continuously into the syringe until only a small amount of sample is left in the syringe.

Eagle Microsystems Handheld Density Meter (SG Pro) — 4.9.7.3 Configure the automatic export of results diagram (page 43)
Fig. 28 · 4.9.7.3 Configure the automatic export of results

The sample flows with a laminar flow into the

measuring cell. 5 Leave the syringe in the connector. 6 Make sure there are no bubbles in the measuring cell.

The measurement starts. 1 When the background of the measuring ribbon

changes color, the measurement is finished. The measured value is saved in the results.

Operation 41 Eagle Eye Power Solutions - SG-Pro

Page 44 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 5.7.1 Configure the method
Page 44 · click to enlarge
Page 44

Drain the measuring cell

1 Pull the plunger continuously out of the syringe. The sample is pulled back into the syringe and

Eagle Microsystems Handheld Density Meter (SG Pro) — 4.9.7.3 Configure the automatic export of results diagram (page 44)
Fig. 29 · 4.9.7.3 Configure the automatic export of results

the measuring cell fills with air. 2 Remove the syringe and empty it into a suitable waste beaker.

5.7 Example: Density determination of cold or hot samples

To achieve good results, the measuring cell and the sample must have the same temperature. If you fill the measuring cell with a sample that is more than 5 °C (9 °F) above or below the temperature displayed on the home screen, this condition is not met. Because the density meter has no active temperature control, you must wait until the temperature of the measuring cell and the temperature of the sample are the same, before you make the measurement. The following chapters show you how to configure and perform a density determination of deionized water that is more than 5 °C (9 °F) above or below the temperature displayed on the home screen.

Summary of the configuration

  • Unguided density determination
  • Filling and draining with the sample pump

You can find more information about the configuration of a method in the chapter [Configure density deter- minations } Page 21].

5.7.1 Configure the method

  • The density unit is set to [g/cm 3 ].
  • Confirmation of results in Settings is activated.
  • The home screen is open. 1 Move the navigation joystick to the left.

The Methods window opens.

2 Press the New key.

The Method settings window opens.

3 Navigate to Method ID and enter a name with up to 25 characters.

The method is listed in the Methods window.

5 Change the parameter settings to the values shown in the following table.

Workflow Manual Users define their individual measuring workflows.

Sample ID entry Variable Users need to enter the identification before they start the measurement.

Fill speed High The pump fills the measuring cell with 80 % of the maximum pump speed.

Measurement reliability Medium The result is saved as soon as the measured temperature value stays within 0.4 °C (0.72 °F) for 10 s.

42 Operation Eagle Eye Power Solutions - SG-Pro

Parameter Setting Explanation

Page 45 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 5.7.2 Perform the density determination
Page 45 · click to enlarge
Page 45

Calculation > Category Density A density is calculated.

d The density is not compensated to a certain temperature.

Calculation > Result limits

None The density meter does not evaluate if the result is within a defined range.

Clean Drain Users are prompted to drain the measuring cell.

Printout None The result is not printed.

Activate and deactivate the confirmation of results } Page 24

5.7.2 Perform the density determination

  • The density meter is set up to work with the sample pump.
  • The piston is in the lowermost position.
  • The home screen is open. 1 Move the navigation joystick to the left.

2 Navigate to the method you want to use and press the Start key. The home screen opens.

3 To enter a Sample ID, move the navigation joystick to the right.

The Sample ID window opens. 4 Enter the identification.

Rinse the measuring cell

1 Immerse the filling tube into the sample (1). 2 Move the navigation joystick upwards and let go.

Eagle Microsystems Handheld Density Meter (SG Pro) — 5.7.2 Perform the density determination diagram (page 45)
Fig. 30 · 5.7.2 Perform the density determination

The sample pump starts and stops when the

piston is in the uppermost position. 3 Hold the filling tube over the waste beaker (2). 4 Move the navigation joystick downwards and let go.

Eagle Microsystems Handheld Density Meter (SG Pro) — 5.7.2 Perform the density determination diagram (page 45)
Fig. 31 · 5.7.2 Perform the density determination

The sample pump starts and stops when the

piston is in the lowermost position. 5 Clean the end of the filling tube with a clean tissue. 6 Repeat the steps once.

Operation 43 Eagle Eye Power Solutions - SG-Pro

Show 4 more labels / short text on this page
Parameter Setting Explanation · Calculation > Calcu- lation · See also · Start the method
Page 46 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 5.8 Read and write information with the RFID reader (DensitoPro only)
Page 46 · click to enlarge
Page 46

Fill the measuring cell and measure the density

1 Immerse the filling tube into the sample (1). 2 Move the navigation joystick upwards and let go.

The sample pump starts and stops when the

piston is in the uppermost position. 3 Make sure there are no bubbles in the measuring cell.

Eagle Microsystems Handheld Density Meter (SG Pro) — 4.9.7.3 Configure the automatic export of results diagram (page 46)
Fig. 32 · 4.9.7.3 Configure the automatic export of results

4 Press the Start key. The measurement starts.

When the background of the measuring ribbon

changes color, the measurement is finished. The measured value is saved in the results.

5 Press the OK key. 6 Wait for about 20 seconds and press the Start key 1 again. The sample that is still in the measuring cell is

measured again.

7 Tap Results and compare the last two results.

8 Tap Back. 9 If the results are not the same, continue to measure the same sample until the temperatures and the results of two measurements in a row are the same.

Drain the measuring cell

1 Hold the filling tube over the waste beaker (1). 2 Move the navigation joystick downwards and let go.

Eagle Microsystems Handheld Density Meter (SG Pro) — 4.9.7.3 Configure the automatic export of results diagram (page 46)
Fig. 33 · 4.9.7.3 Configure the automatic export of results

The sample pump starts and stops when the

piston is in the lowermost position. 3 Clean the end of the filling tube with a clean tissue.

5.8 Read and write information with the RFID reader (DensitoPro only)

With the RFID reader, you can read or write identifications for samples or standards to Smart Tags. The RFID reader only recognizes the following characters.

  • A-Z
  • a-z
  • 0-9

Metal containers interfere with reading and writing to Smart Tags. For this reason you need to place the Smart Tag (1) on a SmartSample film (2).

44 Operation Eagle Eye Power Solutions - SG-Pro

Page 47 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 5.9 Read information with the barcode reader (DensitoPro only)
Page 47 · click to enlarge
Page 47

See also Miscellaneous accessories } Page 66

5.8.1 Write information to a Smart Tag

  • The RFID reader is activated.
  • METTLER TOLEDO Smart Tags are available. 1 Press the Menu key.

2 Navigate to Settings > RFID / Barcode >Write ID to RFID tag. The icon is displayed in the status ribbon. 3 Enter the identification. 4 Hold the RFID reader over a Smart Tag and press the Write key. An audio signal is played and the notification

Eagle Microsystems Handheld Density Meter (SG Pro) — 5.8.1 Write information to a Smart Tag diagram (page 47)
Fig. 34 · 5.8.1 Write information to a Smart Tag

RFID tag is written is displayed. The identification is saved on the Smart Tag.

Activate and deactivate the RFID reader (DensitoPro

only) } Page 13

5.8.2 Read information from a Smart Tag

  • Enter the identification of a sample or a standard
  • The RFID reader is activated.
  • The parameter Sample ID entry or Standard ID entry is set to Variable . −When you are prompted to enter an identification,
Eagle Microsystems Handheld Density Meter (SG Pro) — 5.8.2 Read information from a Smart Tag diagram (page 47)
Fig. 35 · 5.8.2 Read information from a Smart Tag

press the RFID key and hold the RFID reader over the Smart Tag. An audio signal is played and the identification is

saved.

Read information from a Smart Tag ▪ The RFID reader is activated. 1 Press the Menu key.

2 Navigate to Settings > RFID / Barcode > Scan RFID tag.

The notification Scan RFID tag is displayed. 3 Hold the RFID reader over a Smart Tag and wait until an audio signal is played. The read information is displayed.

Eagle Microsystems Handheld Density Meter (SG Pro) — 5.8.2 Read information from a Smart Tag diagram (page 47)
Fig. 36 · 5.8.2 Read information from a Smart Tag

Activate and deactivate the RFID reader (DensitoPro

only) } Page 13 Configure density determinations } Page 21

5.9 Read information with the barcode reader (DensitoPro only)

With the barcode reader, you can read identifications for samples and standards from linear barcodes.

Operation 45 Eagle Eye Power Solutions - SG-Pro

Show 5 more labels / short text on this page
<2 cm · See also · <2 cm · See also · <2 cm
Page 48 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 5.10.3 Export and print results
Page 48 · click to enlarge
Page 48

The barcode reader only recognizes the following characters.

  • A-Z
  • a-z
  • 0-9
  • Procedure
  • The barcode reader is activated.
  • The parameter Sample ID entry or Standard ID entry is set to Variable .

− When you are prompted to enter an identification, press the Barcode key and hold the barcode reader

over the barcode. An audio signal is played and the identification is saved.

See also Activate and deactivate the barcode reader (DensitoPro only) } Page 13 Configure density determinations } Page 21

5.10 View, export, print and delete results

5.10.1 View results

  • The home screen is open. 1 Press the Results key. A list with all results is displayed. 2 To view an individual result, navigate to the result and move the navigation joystick to the right.

5.10.2 Delete all results

You can only delete all results but not individual results or a subgroup of results. ▪ The home screen is open. 1 Press the Results key.

2 Press the Delete key.

The message Do you want to delete all results? is displayed.

3 Press the Yes key. An empty list is displayed.

5.10.3 Export and print results

NOTICE

Loss of data! A maximum of 1100 results is saved in the density meter. When the result list contains 1100 results and a new results is saved, the oldest result is overwritten. − Print or export the results before the maximum number of results is reached.

You can export or print results if you have the required accessories.

  • To export results you need the EasyDirect Density & Refractometry software.
  • To print results, you need a compatible USB printer or bluetooth printer.

46 Operation Eagle Eye Power Solutions - SG-Pro

Page 49 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 5.10.3.2 Print results
Page 49 · click to enlarge
Page 49

See also Accessories } Page 65 Connect to EasyDirect Density & Refractometry } Page 17 Connect a printer } Page 15

5.10.3.1 Export results to EasyDirect Density & Refractometry

  • Export an individual result
  • EasyDirect Density & Refractometry is connected.
  • An individual result is displayed. 1 Press the Report key.

2 Navigate to Option and select EasyDirect.

4 Press the Start key. The export is completed and the result is displayed.

  • Export all results or a subset of results
  • EasyDirect Density & Refractometry is connected.
  • The result list is displayed. 1 Press the Report key.

2 Navigate to Option and select EasyDirect.

4 To export all results, deactivate the filter. 5 To export a subset of results, configure the filter.

6 Press the Start key. The export is completed and the result list is displayed.

See also Configure the filter } Page 48 Connect to EasyDirect Density & Refractometry } Page 17

5.10.3.2 Print results

The Printout parameter defines whether a result is automatically printed when it is saved and which infor- mation is printed.

Information Methods Test Adjustment

Date / Time Short/Long Short/Long Short/Long

Method ID Short/Long – –

Sample ID Short/Long – –

Application name with standard

– Short/Long Short/Long

Standard ID – Short/Long Short/Long

Status Short/Long Short/Long –

Calculation Short/Long – –

Density Short/Long Short/Long Short/Long

Dev. d – Short/Long Short/Long

Temp. Short/Long Short/Long Short/Long

d nominal – Short/Long Short/Long

SNR Long Long Long

Operation 47 Eagle Eye Power Solutions - SG-Pro

Page 50 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 5.10.3.3 Configure the filter
Page 50 · click to enlarge
Page 50

Information Methods Test Adjustment

Additional settings like the compensation temperature and the temperature-compen-

  • Print an individual result
  • A printer is connected.
  • An individual result is displayed. 1 Press the Report key.

2 Navigate to Option and select Print.

4 Navigate to Printout and set it to Short or Long as needed.

The result is printed.

  • Print all results or a subset of results
  • A printer is connected.
  • The result list is displayed. 1 Press the Report key.

2 Navigate to Option and select Print.

4 Navigate to Printout and set it to Short or Long as needed.

6 To print all results, deactivate the filter. 7 To print a subset of results, configure the filter.

8 Press the Start key.

The results are printed.

See also Configure the filter } Page 48 Connect a printer } Page 15

5.10.3.3 Configure the filter

If you activate the filter, you can print or export a subset of the results. The following filters are available.

  • Off : Filter is deactivated and all results are printed or exported.
  • Today : Only the results from current day are printed or exported.
  • Range : Only the results within the set date range are printed or exported.

1 Press the Menu key.

2 Navigate to Results > Report > Filter.

3 Select the desired Filter type.

5 If you selected Range, define the values for Start date and End date.

See also Export results to EasyDirect Density & Refractometry } Page 47 Print results } Page 47

48 Operation Eagle Eye Power Solutions - SG-Pro

Long Long Long · sation coefficient α · Procedure

Page 51 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 6.2.1 Clean the housing
Page 51 · click to enlarge
Page 51

6 Maintenance

In this chapter you find descriptions of the maintenance tasks you should perform on your density meter. Any other maintenance tasks need to be performed by a service technician that has been qualified by EEPS.

If you experience problems with your density meter, contact your authorized EEPS dealer or service representative.

EEPS recommends that a preventive maintenance and calibration certification is done at least once a year through your authorized EEPS dealer or service representative.

u www.mt.com/contact

6.1 Maintenance schedule

If the standard operating procedures of your company require other maintenance intervals, use the intervals listed in the standard operating procedures.

Daily Clean the measuring cell at the end of the work day.

[Clean the measuring cell } Page 50]

Perform a test with deionized water. [Check the measurement accuracy } Page 51]

Activate and deactivate a cleaning reminder } Page 20

6.2 Clean the density meter

NOTICE

Danger of damage to the density meter due to inappropriate cleaning methods! Inappropriate cleaning agents can damage the housing or other parts of the density meter. If liquids enter the housing they can damage the density meter. 1 Make sure the cleaning agent is compatible with the material of the part you want to clean. 2 Make sure that no liquid enters the interior of the density meter.

If you have questions about the compatibility of cleaning agents, contact your authorized EEPS dealer or service representative.

u www.mt.com/contact

6.2.1 Clean the housing

EEPS recommends the following cleaning agents:

  • Water
  • Water with a mild detergent

Procedure ▪ The density meter is shut down. −Wipe the housing with a cloth moistened with the cleaning agent.

Maintenance 49 Eagle Eye Power Solutions - SG-Pro

Frequency Task See · See also

Page 52 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 6.2.2.1 Clean using the sample pump
Page 52 · click to enlarge
Page 52

6.2.2 Clean the measuring cell

Because the measuring cell always contains residue from the last sample, there is a risk that the measuring cell can be damaged by the sample. To prevent such damages, EEPS recommends that you clean the measuring cell at the end of each sample series.

To clean the measuring cell, you need to rinse the measuring cell with a cleaning solution. The cleaning solution must have the following properties.

  • Does not chemically interact with the material of the measuring cell.
  • Dissolves the sample that was just measured.
  • Evaporates without leaving incrustations. EEPS recommends the following cleaning agents:

Acids, concentrated Water (flush the measuring cell with plenty of water to remove the heat from the reaction of water and acid) Alkaline, concentrated 0.3…0.5 % deconex solution

Samples with fats or oily compoents 0.3…0.5 % deconex solution

Petrochemical samples Edible oils and fats

Toluene, xylene or petrol ether mixtures

Activate and deactivate a cleaning reminder } Page 20

6.2.2.1 Clean using the sample pump

  • The density meter is set up to work with the sample pump.
  • The cleaning solution is defined and available.
  • The piston is in the lowermost position. 1 Immerse the filling tube in the cleaning solution (1). 2 Move the navigation joystick upwards and let go.
Eagle Microsystems Handheld Density Meter (SG Pro) — 6.2.2.1 Clean using the sample pump diagram (page 52)
Fig. 37 · 6.2.2.1 Clean using the sample pump

The sample pump starts and stops when the

piston is in the uppermost position. 3 Lift the filling tube out of the cleaning solution and hold it over the waste beaker (2). 4 Move the navigation joystick downwards and let go.

Eagle Microsystems Handheld Density Meter (SG Pro) — 6.2.2.1 Clean using the sample pump diagram (page 52)
Fig. 38 · 6.2.2.1 Clean using the sample pump

The sample pump starts and stops when the

piston is in the lowermost position. 5 Clean the end of the filling tube with a clean tissue. 6 Repeat the steps until all residue from the sample is removed. 7 Move the navigation joystick upwards and let go.

The sample pump fills the measuring cell with

air. 8 Move the navigation joystick downwards and let go. Most of the cleaning solution is removed from the measuring cell. 9 Clean the end of the filling tube with a clean tissue.

50 Maintenance Eagle Eye Power Solutions - SG-Pro

Show 4 more labels / short text on this page
Sample Cleaning solution · Water Water based · Deionized water · See also
Page 53 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 6.3.1 Example: Test with water
Page 53 · click to enlarge
Page 53

6.2.2.2 Clean using a syringe

  • The density meter is set up to work with a syringe.
  • The cleaning solution is defined and available.
  • The measuring cell is drained. 1 Fill a syringe with the cleaning solution. 2 Insert the syringe into the connector for the syringe.
Eagle Microsystems Handheld Density Meter (SG Pro) — 6.2.2.2 Clean using a syringe diagram (page 53)
Fig. 39 · 6.2.2.2 Clean using a syringe

3 Hold the filling tube over the waste beaker (1). 4 Press the plunger continuously into the syringe until the syringe is empty.

Eagle Microsystems Handheld Density Meter (SG Pro) — 6.2.2.2 Clean using a syringe diagram (page 53)
Fig. 40 · 6.2.2.2 Clean using a syringe

5 Pull the plunger continuously out of the syringe.

Eagle Microsystems Handheld Density Meter (SG Pro) — 6.2.2.2 Clean using a syringe diagram (page 53)
Fig. 41 · 6.2.2.2 Clean using a syringe

The cleaning solution is pulled back into the

syringe and the measuring cell fills with air. 6 Remove the syringe and empty it into a suitable waste beaker. 7 Repeat the steps until all residue from the sample is removed.

6.3 Check the measurement accuracy

To check the measurement accuracy of the density meter, you need to configure and perform a test.

You can find more information about the configuration of a test in the chapter [Configure density determi- nations } Page 21].

6.3.1 Example: Test with water

The following chapters show you how to configure and perform a guided test with deionized water at room temperature (23 °C or 73.4 °C).

Maintenance 51 Eagle Eye Power Solutions - SG-Pro

Page 54 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 6.3.1.2 Perform the test
Page 54 · click to enlarge
Page 54

6.3.1.1 Configure the test

  • The density unit is set to [g/cm 3 ].
  • Confirmation of results in Settings is activated. 1 Press the Menu key.

2 Navigate to Test.

The Test window opens.

3 Change the parameter settings to the values shown in the following table.

Workflow Guided The density meter prompts users to fill and drain the measuring cell.

Standard type Water The density meter uses predefined values for the water measurement.

Standard ID entry Variable The density meter prompts users to enter the identifi- cation before the density is measured.

Fill cycle 3 The density meter prompts users to fill and drain the measuring cell twice and then fill the measuring cell for the measurement.

Fill speed High The pump fills the measuring cell with 80 % of the maximum pump speed.

Tolerance d 0.0002 g/cm3 The result is marked as passed if it falls within the following range.

d nominal ± Tolerance d

Clean Drain Users are prompted to drain the measuring cell.

Printout None The result is not printed.

See also Activate and deactivate the confirmation of results } Page 24 Change the density unit } Page 18

6.3.1.2 Perform the test

  • The density meter is set up to work with the sample pump.
  • The measuring cell has been cleaned.
  • The piston is in the lowermost position. 1 Press the Menu key.

2 Navigate to Test.

The Test window opens.

3 Press the Start key. The home screen opens.

4 Press the Start key.

The Standard ID window opens.

5 Enter the identification.

Rinse the measuring cell

1 Immerse the filling tube in the water. 2 Press the OK key. The sample pump starts and stops when the piston is in the uppermost position.

52 Maintenance Eagle Eye Power Solutions - SG-Pro

Parameter Setting Explanation · Start the test

Page 55 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 6.4 Replace the battery
Page 55 · click to enlarge
Page 55

3 Hold the filling tube over the waste beaker.

4 Press the OK key. The sample pump starts and stops when the piston is in the lowermost position. 5 Clean the end of the filling tube with a clean tissue. 6 Repeat the steps once.

Fill the measuring cell and measure the density 1 Immerse the filling tube in the water.

2 Press the OK key. The sample pump starts and stops when the piston is in the uppermost position. 3 Make sure there are no bubbles in the measuring cell. The measurement starts. When the background of the measuring ribbon changes color, the measurement is finished. The measured value is saved in the results.

Drain the measuring cell

1 Hold the filling tube over the waste beaker.

2 Press the OK key. The sample pump starts and stops when the piston is in the lowermost position. 3 Clean the end of the filling tube with a clean tissue.

6.3.2 Measures if a test fails

1 Check if the correct standard has been used, and if needed repeat the test with the correct standard. 2 If the test continues to fail, clean the measuring cell with a cleaning solution that dissolves your samples and dissolves in the standard. 3 Repeat the test. 4 If the test continues to fail, repeat the test twice and compare all three results. 5 If the results are not the same, clean the measuring cell and perform tests until the test passes or the results of three consecutive tests are the same. 6 If the test continues to fail and the results of three consecutive tests are the same, restore the factory adjustment data and repeat the test. 7 If the test continues to fail, perform an adjustment.

See also Clean the measuring cell } Page 50 Restore the factory adjustment data } Page 58 Adjust the measurement accuracy } Page 58

6.4 Replace the battery

The density meter works with a rechargeable 2400 mAh Li-Ion battery that you can order from EEPS.

Maintenance 53 Eagle Eye Power Solutions - SG-Pro

Page 56 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 6.6 Replace the sample pump
Page 56 · click to enlarge
Page 56

1 Press on the lower part (1) of the battery cover and push the battery cover up. 2 Pull the upper part of the battery out of the battery compartment and lift the battery out.

Eagle Microsystems Handheld Density Meter (SG Pro) — 4.9.7.3 Configure the automatic export of results diagram (page 56)
Fig. 42 · 4.9.7.3 Configure the automatic export of results
Eagle Microsystems Handheld Density Meter (SG Pro) — 4.9.7.3 Configure the automatic export of results diagram (page 56)
Fig. 43 · 4.9.7.3 Configure the automatic export of results

3 Insert the lower part of the battery at an angle into the battery compartment. 4 Push the upper part of the battery into the battery compartment. 5 Slide the battery cover from above onto the density meter.

Eagle Microsystems Handheld Density Meter (SG Pro) — 4.9.7.3 Configure the automatic export of results diagram (page 56)
Fig. 44 · 4.9.7.3 Configure the automatic export of results
Eagle Microsystems Handheld Density Meter (SG Pro) — 4.9.7.3 Configure the automatic export of results diagram (page 56)
Fig. 45 · 4.9.7.3 Configure the automatic export of results

See also Accessories } Page 65 Technical data } Page 62

6.5 Replace the filling tube

1 Unscrew the filling tube by hand and remove it. 2 Install the new filling tube.

Eagle Microsystems Handheld Density Meter (SG Pro) — 6.5 Replace the filling tube diagram (page 56)
Fig. 46 · 6.5 Replace the filling tube

See also Accessories } Page 65 Install the filling tube } Page 13

6.6 Replace the sample pump

Remove the sample pump

  • The measuring cell is drained. 1 Move the navigation joystick upwards and let go. The sample pump starts and stops when the piston is in the uppermost position. 2 Shut down the density meter.

54 Maintenance Eagle Eye Power Solutions - SG-Pro

Page 57 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 6.9 Ship the density meter
Page 57 · click to enlarge
Page 57

3 Press on the grip recesses (4) of the cover and lift the lower part of the cover out of the housing. 4 Pull the top of the cover out of the housing. 5 Lift the piston (1) out of the guide system. 6 Pull the cylinder (2) up and out of its guide system (3).

Eagle Microsystems Handheld Density Meter (SG Pro) — 4.9.7.3 Configure the automatic export of results diagram (page 57)
Fig. 47 · 4.9.7.3 Configure the automatic export of results
Eagle Microsystems Handheld Density Meter (SG Pro) — 4.9.7.3 Configure the automatic export of results diagram (page 57)
Fig. 48 · 4.9.7.3 Configure the automatic export of results

Install the sample pump

1 Insert the cylinder (3) of the sample pump in its guide system (4). 2 Insert the piston (1) in its guide system (2). 3 To check if the sample pump works, move the navigation joystick downwards and let go.

Eagle Microsystems Handheld Density Meter (SG Pro) — 4.9.7.3 Configure the automatic export of results diagram (page 57)
Fig. 49 · 4.9.7.3 Configure the automatic export of results

The sample pump starts and stops when the

piston is in the lowermost position. 4 Reinstall the cover of the sample pump.

See also Accessories } Page 65

6.7 Update the firmware

It is possible to update the firmware of your density meter. To update the firmware of the density meter, contact an authorized EEPS dealer or service representative.

u www.mt.com/contact

6.8 Prepare the density meter for storage

The density meter can be stored at temperatures of -20…+50 °C (-4…+122 °F). But the lifespan of the battery is reduced when it is stored at high temperatures. 1 Clean the measuring cell. 2 Shut down the density meter. 3 Clean the housing. 4 Store the density meter in a dry and clean place.

See also Clean the density meter } Page 49 Technical data } Page 62

6.9 Ship the density meter

If you have questions about transporting your density meter, contact your authorized EEPS dealer or service representative.

u www.mt.com/contact

1 Shut down the density meter.

Maintenance 55 Eagle Eye Power Solutions - SG-Pro

Page 58 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 6.10 Dispose of the density meter
Page 58 · click to enlarge
Page 58

2 Clean the density meter. 3 Ship the density meter using the original packaging.

6.10 Dispose of the density meter

In conformance with the European Directive 2012/19/EU on Waste Electrical and Electronic Equipment (WEEE) this device may not be disposed of in domestic waste. This also applies to countries outside the EU, per their specific requirements.

Please dispose of this product in accordance with local regulations at the collecting point specified for electrical and electronic equipment. If you have any questions, please contact the responsible authority or the distributor from which you purchased this device. Should this device be passed on to other parties (for private or professional use), the content of this regulation must also be related.

Thank you for your contribution to environmental protection.

56 Maintenance Eagle Eye Power Solutions - SG-Pro

Page 59 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 7 Troubleshooting
Page 59 · click to enlarge
Page 59

7 Troubleshooting

7.1 List of errors and problems

Problem Possible cause Measure

The sample pump does not aspirate any liquid.

The locking screw is not installed. − Install the locking screw.

The locking screw is not tight. − Tighten the locking screw. The measuring cell is not clean. See [Measures if a test fails } Page 53].

Test of measurement accuracy fails.

Mismatch between measured and configured standard

See [Measures if a test fails } Page 53].

The tolerance range is too small. See [Configure the tolerance for a

test } Page 33].

The density meter is not adjusted correctly.

See [Measures if a test fails } Page 53]. 1 Hold the filling tube over the waste beaker. 2 Move the navigation joystick downwards and let go. The sample pump starts

The message Drain cell over waste beaker. is displayed when you try to start a method.

The piston is not in the lowermost position.

and stops when the piston is in the lowermost position. 3 Clean the end of the filling tube with a clean tissue.

There are bubbles in the measuring cell.

Fill speed is too high. 1 If you use the sample pump, lower the setting for fill speed. See [Define the pump speed } Page 23]. 2 If you use a syringe, press the syringe plunger more slowly into the syringe.

The liquid contains dissolved gases.

− Degas the sample before you

fill it into the measuring cell.

Connector for syringe or the filling tube is not tight.

1 Tighten the connector of the filling tube. 2 Tighten the locking screw or the connector for the syringe.

Troubleshooting 57 Eagle Eye Power Solutions - SG-Pro

Page 60 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 7.3 Adjust the measurement accuracy
Page 60 · click to enlarge
Page 60

Problem Possible cause Measure

The message Pump position unknown. is displayed.

The saved pump position was lost because the battery was removed too long.

The density meter is set up to work with a syringe.

− You can ignore the message

and continue to work with the density meter.

The density meter is set up to work with the sample pump. 1 Hold the filling tube over the waste beaker. 2 Move the navigation joystick downwards and let go. The sample pump starts

and stops when the piston is in the lowermost position. 3 Clean the end of the filling tube with a clean tissue.

"--" is displayed, instead of a value for Alpha*1000.

The calculated value is out of range.

− Check and correct the density

values and the values for the compensation temperatures you entered.

"--" is displayed, instead of a The measured value or the result is 1 Check that the unit defined in measured value or a result. out of range. the method matches the sample you want to measure. 2 If you use a user-defined conversion, check that all the settings are correct.

7.2 Restore the factory adjustment data

If you restore the factory adjustment data, the current adjustment data is overwritten with the adjustment data that were determined when the density meter was adjusted during production.

1 Press the Menu key.

2 Navigate to Maintenance & Service > Reset to factory adjust..

The message Do you want to reset to factory adjustment? opens.

3 To overwrite the current adjustment data with the factory adjustment data, press Yes.

The factory adjustment data are restored.

7.3 Adjust the measurement accuracy

NOTICE

The following chapters show you, how to configure and perform an exemplary adjustment with water.

You can find more information about the configuration of an adjustment in the chapter [Configure density determinations } Page 21].

58 Troubleshooting Eagle Eye Power Solutions - SG-Pro

Page 61 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 7.3.1.2 Perform the adjustment
Page 61 · click to enlarge
Page 61

See also Measures if a test fails } Page 53

7.3.1 Example: Adjustment with water

The following chapters show you how to configure and perform a guided adjustment with deionized water at room temperature (23 °C or 73.4 °C).

7.3.1.1 Configure the adjustment

  • Confirmation of results in Settings is activated. 1 Press the Menu key.

2 Navigate to Adjustment.

The Adjustment window opens.

3 Change the parameter settings to the values shown in the following table.

Workflow Guided The density meter prompts users to fill and drain the measuring cell. Standard type Water The density meter uses predefined values for the water measurement.

Standard ID entry Variable The density meter prompts users to enter the identifi- cation before the density is measured. Fill cycle 3 The density meter prompts users to fill and drain the measuring cell twice and then fill the measuring cell for the measurement.

Fill speed High The pump fills the measuring cell with 80 % of the maximum pump speed. Clean Drain Users are prompted to drain the measuring cell.

Printout None The result is not printed.

7.3.1.2 Perform the adjustment

  • The density meter is set up to work with the sample pump.
  • The measuring cell has been cleaned.
  • The piston is in the lowermost position. 1 Press the Menu key.

2 Navigate to Adjustment.

The Adjustment window opens.

3 Press the Start key. The home screen opens.

4 Press the Start key.

The Standard ID window opens.

5 Enter the identification.

Troubleshooting 59 Eagle Eye Power Solutions - SG-Pro

Parameter Setting Explanation · Start the adjustment

Page 62 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 4.9.7.3 Configure the automatic export of results
Page 62 · click to enlarge
Page 62

Rinse the measuring cell

1 Immerse the filling tube in the water (1).

Eagle Microsystems Handheld Density Meter (SG Pro) — 4.9.7.3 Configure the automatic export of results diagram (page 62)
Fig. 50 · 4.9.7.3 Configure the automatic export of results

The sample pump starts and stops when the

piston is in the uppermost position. 3 Hold the filling tube over the waste beaker (2).

Eagle Microsystems Handheld Density Meter (SG Pro) — 4.9.7.3 Configure the automatic export of results diagram (page 62)
Fig. 51 · 4.9.7.3 Configure the automatic export of results

The sample pump starts and stops when the

piston is in the lowermost position. 5 Clean the end of the filling tube with a clean tissue. 6 Repeat the steps once.

Fill the measuring cell and measure the density 1 Immerse the filling tube in the water (1).

The sample pump starts and stops when the

piston is in the uppermost position. 3 Make sure there are no bubbles in the measuring cell. The measurement starts.

Eagle Microsystems Handheld Density Meter (SG Pro) — 4.9.7.3 Configure the automatic export of results diagram (page 62)
Fig. 52 · 4.9.7.3 Configure the automatic export of results

When the background of the measuring ribbon

changes color, the measurement is finished. The measured value is saved in the results. 4 Accept or reject the adjustment.

Drain the measuring cell

1 Hold the filling tube over the waste beaker (1).

Eagle Microsystems Handheld Density Meter (SG Pro) — 4.9.7.3 Configure the automatic export of results diagram (page 62)
Fig. 53 · 4.9.7.3 Configure the automatic export of results

The sample pump starts and stops when the

piston is in the lowermost position. 3 Clean the end of the filling tube with a clean tissue.

Activate and deactivate the confirmation of

results } Page 24

60 Troubleshooting Eagle Eye Power Solutions - SG-Pro

See also

Page 63 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 7.5 View the firmware version and other system information
Page 63 · click to enlarge
Page 63

7.4 Restore the factory settings

NOTICE

Loss of data With a factory reset all user-specific settings are set to the default values and all saved data (e.g. sample IDs, results, Method IDs and adjustment data) are deleted.

−Document the data you need before you restore the factory settings.

1 Press the Menu key.

2 Navigate to Maintenance & Service > Reset to factory settings.

The dialog Do you want to reset all values to factory settings? opens.

3 To reset the density meter to factory settings, press the Yes key.

The density meter restarts.

7.5 View the firmware version and other system information

  • The home screen is open. 1 Press the Menu key.

2 Navigate to Information.

The firmware version and other system information are displayed.

Instrument type Shows if the density meter is a SG-Pro or DensitoPro.

SNR instrument Shows the serial number of the instrument.

Firmware version Shows the firmware version of the density meter. Licenses Opens a window with information on the End User License Agreement, third party licenses and open source attribution files.

Troubleshooting 61 Eagle Eye Power Solutions - SG-Pro

Page 64 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 8 Technical data
Page 64 · click to enlarge
Page 64

8 Technical data

8.1 Density meter

Power rating instrument Input values 5 V DC, 1.0 A

Connector type USB C Power rating AC adapter Input values 100…240 V AC, 0.3 A

Input frequency 50 - 60 Hz Output values 5 V DC, 1.0 A

Battery Type Rechargeable lithium-ion battery Capacity 2400 mAh Temperature range for charging 0…40 °C Dimensions Width 56 mm

Length 288 mm Depth 80 mm

Weight 355 g Screen Technology Color screen Size 2.4"

Resolution 240 x 320 pixels USB type A Data rate USB 2.0, host fullspeed

Output voltage 5 V DC ±5 % Max. output current 250 mA

USB type C Data rate USB 2.0, device fullspeed

Input voltage 5 V DC ±10 % Max. input current 400 mA, max. 1 A with AC adapter

RFID reader (DensitoPro only)

Compatibility METTLER TOLEDO Smart Tags

Barcode reader (DensitoPro only)

Materials Housing PBT (polybutylene terephthalate) Measuring cell Borosilicate glass

Parts in contact when working with sample pump

PTFE (polytetrafluorethylene) PVDF (polyvinylidene difluoride) EPDM (ethylene propylene diene monomer (M- class) rubber) FFKM (perfluoro-elastomers) Borosilicate glass Filling tube PTFE (polytetrafluorethylene)

Connector for filling tube or syringe

Ambient temperature (SG-Pro) -10…+50 °C Ambient temperature (DensitoPro) 0…+50 °C

Relative humidity Max. 80 % (non-condensing) at 31 °C, linear fall to 50 % at 40 °C

62 Technical data Eagle Eye Power Solutions - SG-Pro

Show 5 more labels / short text on this page
Characteristic Value · Format Linear barcodes · PP (polypropylene) · Ambient conditions · Characteristic Value
Page 65 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 8.2 Measurement
Page 65 · click to enlarge
Page 65

Altitude Up to 2000 m above sea level

Use In interior spaces

Instrument -20 °C…+50 °C

Battery Storage for up to 1 month -20 °C…+60 °C

Storage for up to 3 months -20 °C…+45 °C

Storage for up to 1 year -20 °C…+30 °C

Directives, standards and REACH regulation

Directives and standards complied with are listed on the declaration of conformity. SVHC candidate substances according to REACH (Article 33)

Material CAS No.

PZT (Lead Zirconate Titanate) 12626-81-2

8.2 Measurement

Density Measuring range 0.000…3.000 g/cm3

Accuracy 1) ± 0.001 g/cm3

Repeatability ± 0.0005 g/cm3 Resolution 0.0001 g/cm3

Temperature Temperature range for sample 0…50 °C

Resolution 0.1 °C

Accuracy ±0.2 °C Viscosity Using sample pump 0...100 mPa*s Using syringe 0...1000 mPa*s

Volumes Sample pump 5 mL Min. sample volume 2 mL

Temperature compensation Automatic or by using a user-defined temperature correction coefficient

Application Max. number of methods 30

1. For viscosity range of 0…200 mPa*s

Ethanol (d) % w/w Range 0.0. 100.0 % w/w

Accuracy ±1.0 % w/w Ethanol (d) % v/v Range 0.0 % v/v

Accuracy ±1.0 % v/v

Technical data 63 Eagle Eye Power Solutions - SG-Pro

Show 12 more labels / short text on this page
Characteristic Value · Pressure Atmospheric pressure · Overvoltage category II · Pollution degree 2 · IP code IP5X · Storage temperatures · Characteristic Value · 1,3-Propanesultone 1120-71-4 · Characteristic Value · Stored results 1100 · Measurement scales · Characteristic Value
Page 66 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 4.9.7.3 Configure the automatic export of results
Page 66 · click to enlarge
Page 66

Proof (IP) Range 0.0. 175.0 Proof

Accuracy ±1.75 Proof

Proof (US) Range 0.0. 200.0 Proof

Accuracy ±2.0 Proof

Light Baumé Range 10...100 °Bé

Accuracy ±0.4 °Bé

Heavy Baumé Range 0...72 °Bé

Accuracy ±0.1 °Bé

H2SO4 % w/w Temperature range 10…50 °C

Concentration range 0.0. 100.0 % w/w

Accuracy ±1.0 % w/w

H2SO4 % v/v Temperature range 10…50 °C

Concentration range 0.0 % v/v

Accuracy ±1.0 % v/v

Brix (d) % w/w Range 0.0. 85.0 % w/w

Accuracy ±0.3 % w/w

Plato Range 0.0. 85.0 °P

Accuracy ±0.3 °P

International standards and norms

International standards and norms complied with are listed on the internet.

64 Technical data Eagle Eye Power Solutions - SG-Pro

Characteristic Value

Page 67 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 9.1 Sampling
Page 67 · click to enlarge
Page 67

9 Accessories

The following accessories and parts are available for the density meters SG-Pro and DensitoPro.

All parts are specified with their order number and quantity in cases where more than one part is included. When ordering, some parts are only available in a minimum order quantity. In such cases, the corre- sponding minimum order quantity is quoted. Certain parts are only available in specific countries. If you have any questions, contact your authorized EEPS dealer.

9.1 Sampling

Filling Tube 190 mm 30330847

Eagle Microsystems Handheld Density Meter (SG Pro) — 9.1 Sampling diagram (page 67)
Fig. 54 · 9.1 Sampling
  • Tube
  • Connector syringe/filling tube
  • Washer

Filling tube 600 mm 30330848

  • Tube
  • Connector syringe/filling tube
  • Washer

Locking Screw, SG-Pro 30330852

Connector syringe/filling tube 30330853

Eagle Microsystems Handheld Density Meter (SG Pro) — 9.1 Sampling diagram (page 67)
Fig. 55 · 9.1 Sampling

Disposable syringe (100 pcs) 10 mL 51338100

  • Piston
  • Glass cylinder
Eagle Microsystems Handheld Density Meter (SG Pro) — 9.1 Sampling diagram (page 67)
Fig. 56 · 9.1 Sampling

Sample pump cover 30330851

Accessories 65 Eagle Eye Power Solutions - SG-Pro

Sample pump · 30330849

Page 68 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 9.3 Miscellaneous accessories
Page 68 · click to enlarge
Page 68

9.2 Printers

Bluetooth Printer Godex MX20 with Dongle 30330864

Bluetooth Printer Paper (5 rolls) 30330865

Bluetooth Dongle Godex MX20 30330863

Eagle Microsystems Handheld Density Meter (SG Pro) — 9.2 Printers diagram (page 68)
Fig. 57 · 9.2 Printers

Lab equip access. data writer USB-P25/01 11124301

9.3 Miscellaneous accessories

EasyDirect Density & Refractometry PC Software 30451628

Density standards (10 pcs) 6 mL 51325005

Battery Li-Ion 2400mAh 30330855

Eagle Microsystems Handheld Density Meter (SG Pro) — 9.3 Miscellaneous accessories diagram (page 68)
Fig. 58 · 9.3 Miscellaneous accessories

66 Accessories Eagle Eye Power Solutions - SG-Pro

Carrying Case 30330861

Page 69 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 4.9.7.3 Configure the automatic export of results
Page 69 · click to enlarge
Page 69
Eagle Microsystems Handheld Density Meter (SG Pro) — 4.9.7.3 Configure the automatic export of results diagram (page 69)
Fig. 59 · 4.9.7.3 Configure the automatic export of results

Power Supply and Worldwide Adapters 30449255

Eagle Microsystems Handheld Density Meter (SG Pro) — 4.9.7.3 Configure the automatic export of results diagram (page 69)
Fig. 60 · 4.9.7.3 Configure the automatic export of results

Tag SmartSample (10 pcs) 30449268

Tag SmartSample Film for Tanks (10 pcs) 30449269

Accessories 67 Eagle Eye Power Solutions - SG-Pro

Battery cover 30330854 · USB-C Cable 30449253 · Accessories for DensitoPro

Page 70 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — 10.2 Temperature-compensation coefficients α
Page 70 · click to enlarge
Page 70

10 Appendix

10.1 Density of pure water (0…40 °C)

Temperature [°C] Density [g/cm3] Temperature [°C] Density [g/cm3]

0 0.99984

1 0.99990 21 0.99799

2 0.99994 22 0.99777

3 0.99996 23 0.99754

4 0.99997 24 0.99730

5 0.99996 25 0.99705

6 0.99994 26 0.99679

7 0.99990 27 0.99652

8 0.99985 28 0.99624

9 0.99978 29 0.99595

10 0.99970 30 0.99565

11 0.99961 31 0.99534

12 0.99950 32 0.99503

13 0.99938 33 0.99471

14 0.99925 34 0.99438

15 0.99910 35 0.99404

16 0.99894 36 0.99369

17 0.99878 37 0.99333

18 0.99860 38 0.99297

19 0.99841 39 0.99260

20 0.99821 40 0.99222

[Chemical Handbook Fundamental Version, Rev. 3, Table 5.2 (1984)]

10.2 Temperature-compensation coefficients α

Substance Temperature range [°C] α • 103/°C

Water 15…30 0.23

Ethanol 0…30 1.09

m-xylene 0…30 0.99

p-xylene 15…30 1.02

Glycerine 15…30 0.49

Chloroform 0…30 1.26

Carbon tetrachloride 0…30 1.22

Toluene 0…30 1.07

Benzene 6…30 1.21

Methanol 6…30 1.18

Acetone 0…30 1.42

Bromobenzene 0…30 0.91

Cyclohexane 0…30 1.20

Isopropanol 0…30 1.06

n-nonane 0…30 1.08

68 Appendix Eagle Eye Power Solutions - SG-Pro

Page 71 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — API 24
Page 71 · click to enlarge
Page 71

Index

A

Activate 20 Deactivate 20

B

Assemble 12 Change prongs 12 Connect 12 Accessories 65

Deactivate 13 Read 45

Auto rotation 20 Barcode reader 13 Cleaning reminder 20

Install 53 Remove 53 Replace 53

Confirmation of results 24 Password protection 19

Storage temperature 63 Technical data 62 Baumé 24, 26 Beep

RFID reader 13 Sample pump mode 13 Syringe mode 14

Configure 23, 29, 30, 31, 32 Configure example 59

Change 18 Dimming 18 Brix 27 C Mass percent 26

Perform 58 Perform example 59

Proof (IP) 26 Proof (US) 26

Volume percent 26 Alpha

Table 68 Ambient conditions 62

Density unit 18 Keyboard layout 20 Language 20 Method 22 Password 20 Prongs AC adapter 12 Screen brightness 18

American Petroleum Institute 24

API 24

API degree 26 Density 26 Specific gravity 26

AC adapter 12 Filling tube 13 Audio signal

Time 17 Time format 17

Density meter 12 Clean

Index 69 Eagle Eye Power Solutions - SG-Pro

Show 29 more labels / short text on this page
AC adapter · Barcode reader · Activate 13 · Activate · Audio signal 20 · Battery · Activate 20 · Adjustment · Deactivate 20 · Change 22 · Bluetooth printer · Connect 15, 16 · Brightness · Density unit 18 · Alcohol concentration 24 · Celsius 18 · Change · Adjustment 22 · Date 17 · Date format 17 · Configure 29 · Temperature unit 18 · Assemble · Test 22 · Activate 20 · Charge · Deactivate 20 · Auto rotation · Housing 49
Page 72 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — D
Page 72 · click to enlarge
Page 72

Measuring cell 50 Cleaning reminder

D

Activate 20 Deactivate 20 Concentration

Sugar 24 Sulfuric acid 24 User-defined 24, 27

Auto rotation 20 Barcode reader 13 Cleaning reminder 20

Adjustment 22, 23, 29, 30, 31, 32

Confirmation of results 24 Password protection 19

Alpha 29 Drain 23

RFID reader 13 Sample pump 14 Sample pump mode 14 Syringe mode 13

Measurement reliability 24 Method 22, 23, 24, 29, 30, 31, 32 Result limits 32 Rinse 23 Standard 32

Method 22 Results 46 Density 24, 25

Test 22, 23, 29, 30, 31, 32, 33 Tolerance 33 User-guidance 23

Temperature compensated 25 Density determination

Adjustment 59 Guided 36, 39 Sample pump 36, 42 Syringe 39

Test 52 Unguided 42

Overview 8 Ship 55 Shut down 34

Store 55 Unpack 11

Density of pure water

Bluetooth printer 15, 16 EasyDirect Density & Refractometry 17 Godex MX20 15, 16 Power supply 12 USB printer 15, 16

Adjustment 18 Change 18

Automatic 18 Disposal 56 Drain

Install 15 Material 62

Background 36 Configure 23 Guided 39, 42 Sampel pump 44

70 Index Eagle Eye Power Solutions - SG-Pro

Show 27 more labels / short text on this page
Date · Change 17 · Date format · Change 17 · Alcohol 24 · Deactivate · Audio signal 20 · Configure · Delete · nfigure example · Typical phases 36 · Density meter · Charge 12 · Functions 8 · Confirm settings 35 · Confirmation of results · Start up 34 · Activate 24 · Deactivate 24 Connect · AC adapter 12 · Table 68 · Density unit · Test 18 Dimming · Connector for syringe · Contact 5 · Create · Method 21
Page 73 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — G
Page 73 · click to enlarge
Page 73

Syringe 42 Unguided 44

H

E

Clean 49 Material 62

EasyDirect Density & Refractometry

Connect 17 Export 32 End user license agreement 5 Energy saver 18

Number 35 Special character 35 Symbol 35 Text 35

Connector for syringe 15 Filling tube 13 Sample pump 55 Syringe adapter 15

K

F

Key functions ribbon 10 Keyboard

Change layout 20 Overview 35

Restore 61 Fahrenheit 18 Fill

Guided 38, 41 Sample pump 44 Syringe 41 Unguided 44

M

Fill speed 23 Filling tube

Connector for filling tube or syringe 62 Filling tube 62 Housing 62 Measuring cell 62 Sample pump 62 Measurement accuracy

Install 13 Material 62 Remove 54 Replace 54

Adjust 58 Check 51 Measurement reliability

End user license agreement 5 Open source attribution files 5 Third party licenses 5

Configure 24 Measuring cell 9 Clean 50 Material 62 Measuring ribbon 10 Method

Update 55 Version 5, 61

G

Index 71 Eagle Eye Power Solutions - SG-Pro

Show 27 more labels / short text on this page
Housing · Icons 10 · Identification · Sample 30 · Enter · Standard 30 · Install · Battery 53 · Errors · List 57 · Export · Intended use · Results 32, 47 · Factory adjustment · Restore 58 · Factory settings · Language · Background 36 · Change 20 · Maintenance · Schedule 49 Material · Assemble 13 · Fir · mware · Godex MX20 · Connect 15, 16 · Change 22
Page 74 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — R
Page 74 · click to enlarge
Page 74

Configure 23, 24, 29, 30, 31, 32

Sample pump 54 Replace

Delete 22 Method ribbon 10

Battery 53 Filling tube 54 Sample pump 54

N

Factory adjustment 58 Factory settings 61

O

Open source attribution files 5

Factory adjustment 58 Result limits

Functions 8 Instrument 8 Keyboard 35

View 46 RFID reader

P

Change 20 Password protection 19 Activate 19 Deactivate 19

Adjustment 58 Test 51 Perform example

Adjustment 59 Method 38, 40, 43 Test 52 Plato 27

R

Barcode reader 45 RFID reader 45 Remove

Battery 53 Filling tube 54 Export 32, 47

72 Index Eagle Eye Power Solutions - SG-Pro

Show 22 more labels / short text on this page
Create 21 · Reset · Navigate 35 · Number · Enter 35 · Restore · Facotry settings 61 · Overview · Configure 32 Results · Delete 46 · Parts 65 Password · Perform · Power supply · Connect 12 · Print · Results 31, 47 · Printer · Connect 15 · Test 16 · Printout · Read · Print 31, 47
Page 75 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — S
Page 75 · click to enlarge
Page 75

Metal container 44 Read 45 Write 45 Rinse

Configure 23 Guided 38, 40 Sample pump 43 Syringe 40 Unguided 43

S

Signal words 6 Warning symbols 6 Sample

Sample ID ribbon 10 Sample pump

Drain 44 Fill 44

Install 55 Material 62 Remove 54 Replace 54

Index 73 Eagle Eye Power Solutions - SG-Pro

Show 6 more labels / short text on this page
Activate 13 · Deactivate 13 · Background 36 · Safety information · Identification 30 · Deactivate 14
Page 76 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — T
Page 76 · click to enlarge
Page 76

Rinse 43 Set up 13 Sample pump mode

Density meter 55 Sucrose concentration 24 Sugar concentration 24

Sampling ribbon 10 Scope of delivery 11

Plato 27 Sulfuric acid concentration 24

Brightness 18 Dimming 18

Density meter 34 Switch on

Sample pump 13 Syringe 14 Ship

Density meter 34 Symbol 5, 10 Enter 35 Syringe

Density meter 55 Shortcut method list 35 Shut down

Automatic 18 Density meter 34 Signal words 6

Fill 41 Rinse 40

Set up 14 Syringe adapter

Site requirements 6, 62

Metal container 44 Read 45 Write 45 Special character

T

Specific gravity 24, 25 Temperature compensated 25 Standard

Configure 32 Identification 30 Standard delivery 11

Configure 29 Values 68

Sample pump 38 Syringe 40 Unguided 43

Configure 23, 29, 30, 31, 32, 33

Density unit 18 Failed 53

Perform 51 Perform example 52 Printer 16

74 Index Eagle Eye Power Solutions - SG-Pro

Show 38 more labels / short text on this page
Status ribbon 10 · Store · Battery 63 · Activate 13 · Deactivate 14 · Brix 27 · Screen · Auto rotation 20 · Mass percent 27 · Volume percent 27 · Serial number 61 · Switch off · Set up · Drain 42 · Smart Tag · Install 15 · Syringe mode · Activate 14 · Deactivate 13 · Enter 35 · Technical data 62 · Temperature compensated · Density 25 · Specific gravity 25 · Temperature unit · Change 18 · Temperature-compensation coefficient 29 · Start adjustment · Guided 59 · Start method · Test · Guided 38, 40 · Change 22 · Configure example 52 · Start test · Guided 52 · Start up · Density meter 34
Page 77 of the Eagle Microsystems Handheld Density Meter (SG Pro) user manual — W
Page 77 · click to enlarge
Page 77

Enter 35 Third party licenses 5

List of errors 57

U

Temperature 18 Unpack 11

Connect 15, 16 User guidance

W

Index 75 Eagle Eye Power Solutions - SG-Pro

Show 22 more labels / short text on this page
Start guided Text · Time · Change 17 · Time format · Change 17 · Tolerance · Configure 33 Troubleshooting · Turn off · Density meter 34 · Turn on · Density meter 34 · Unit · Density 18 · Update · Firmware 55 · USB printer · Configure 23 · View · Firmware version 61 · Results 46 · Write · RFID reader 45

Ask about the Eagle Microsystems Handheld Density Meter (SG Pro)

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions