Oxymeter De Pouls Digital MANUAL USER Ver0.0 Model Number: AE-B4 Description of General Order Pulse oxygenometer (saturometer) is an electronic device that enables the measurement, in real time and non-invasively, simple, reliable and continuous, of the Oxygen Hemoglobin Pulsate Saturation (SPO2). Oxygen (O2) saturation of the Hemoglobin is obtained by the value (percentage) of the capacity for oxyhemoglobe ("Hb O2"), composed of oxygen combined with hemoglobulin, divided by the total blood-mixed haemoglobin (Hb). In other words, it is the consistency of erythioglobin in blood. This parameter is very important in the area of Respiratory Traffic System. A large number of respiratory infections are likely to cause a reduction in the oxygen saturation rate of haemoglobin present in human blood. In addition, the factor to be followed is also likely to reduce oxygen saturation: Failure of automatic regulation of organic functions. This may be caused by anaesthesia, post-operative trauma or injuries resulting from medical examinations. This situation may cause mild headaches, asthenia and vomiting. It is therefore very important to know the oxygen saturation of plasma hemoglobin in the client, with a view for medical professionals to be able to detect problems upstream. The characteristics of the digital pulse oxymeter (saturometer) are: low consumption, practical use and operation, increased mobility. To obtain a precise diagnosis and to allow for an on-screen display of the Pouls Frequency and Pulse Saturation of Hemoglobin Oxygen, position the photoelectric sensor at one end of one of the patient's fingers. Clinical trials have also demonstrated the high degree of precision and reliability of this type of device. Measurement Principle The Pulse Oxymeter principle is as follows: a mathematical formula was developed from the Beer-Lambert Law using the Absorption Characteristics of the Reduced Hemoglobin Spectrum (non-oxygenated, or "RHb") and Oxyhemoglobin (oxygenized hemoglobe, or HB O2) for red and near infrared areas. The operating principle of the measuring instrument is that a Photoelectric Technology Device d-Oxyhemoglobin inspection is used in conjunction with a Pouls Scanning System and a Recording Technology. The blood colour varies according to its oxygenation, that is, saturation. Thanks to this property, the sensors of devices will be able to determine SPO2. In practice, this instrument is equipped with a "clip-able" sensor at the finger level capable of transmitting signals via two diodes having an wavelength of 660 nm (nanometre) (red) and 940 nm(near infrared). The apparatus will thus identify each pulsatile blood wave and then determine the saturation according to the color of the blood that will be deduced based on absorption of the emitted lights. The light amplitude (the measured signal) is captured by a photoreceptor and the value displays, via electronic processing (circuits and

microprocessor), at the saturometer screen. Operating Principle Diagram 1. Tube of a Red and Infrared Radiation. 2. Reception tube of a Red and Infrared Ray. Precautions for use 1 Before first use, please read this manual carefully. 2 Do not use digital pulse oxygenmeter near an Electro-Surgical Unit ("ESU"), as this device may be disturbed. 3 Touch pulse oxygenmeter shall be positioned to allow for correct pulse capture, the primary objective being reliable determination of a pulsed oxygen saturation value (SP O2). Before determining the pulsed oxygen saturation value, check that nothing compromises the correct pulse capture. 4 Do not use the saturometer in an MRI environment ("Imaging by Magnetic Resonance") or CT ("Digital Tomography"). 5 Do not use the touch pulse oxygenometer in situations requiring alarms. This device is not equipped with alarms and is not intended for continuous monitoring. 6 Do not use the saturometer in an explosive atmosphere. 7 Tactile pulse oxygenometer is intended to help assess the patient's health status. It should be used in addition to other methods for assessing clinical symptoms and signs. 8 In order to better understand the correct positioning of the sensor and check, not only circulation but also so-called "skin sensitivity" of the patient, frequently (every 4 hours) control the area of application of the pulse oxymeter sensor. . 9 Do not sterilize the apparatus with an autoclave, d-oxide or by immerging in a liquid. This device shall not be sterilized. 10 Follow local orders and recycling instructions for the disposal and recycling of this type of instrument, its components (including batteries). 11 This equipment meets the requirements of IEC 60601-1-1-2 (2007) concerning electromagnetic compatibility of "electromedical" systems and/or equipment. However, due to the proliferation of RF transmission equipment and other so-called "electric" nuisances in health and other environments, high degrees may be possible. interference from increased proximity or high power of a source that is the cause of lower performance of this device. 12 Mobile and portable RF communication equipment can cause malfunctions in "medical" electrical devices. "Rx only" (literally, "only on prescription"): "Warning: The federal law restricts the sale of this device to health professionals. »


Inaccurate measurements may be obtained from/d-un/d-1 Significant levels of dysfunctional hemoglobin (carboxyhemoglobin or methaemoglobe); 2 Intravascular colours (green d-indocyanin or blue methylene); 3 Ambient light too strong. If necessary, protect the sensor positioning area; 4 Excessive patient movements; 5 Deficits and interference of electrosurgical apparatus known as "high frequency"; 6 Venous pulsations; 7 Positioning a sensor on an end equipped with tensiometer armband, arterial catheter or intravascular line; 8 Hypotension, severe vasoconstriction, serious anaemia or hypothermia of the patient; 9 Cardiac arrest or stress of the patients; 10 Nail polishes (colorimetric interference) or presence of false nails; 11 Low pulse quality (low infusion); 12 Low haemoglobin levels. Product Properties 1 Use of the product is simple and convenient. 2 It is compact, lightweight (it weighs about 50g with batteries) and easy to carry. 3 The two original AAA batteries can operate continuously for 30 hours. 4 A signal indicates that the batteries are weak. Caution when the battery level is low it could affect the proper operation of the oximeter. 5 The device will turn off automatically when no signal is issued for more than 8 seconds. Target Use The digital oxygen meter is used to measure human hemoglobin oxygen saturation and heart rate by finger. The product can be used in the family, at hospital (which includes clinical use in internal surgery, anaesthesia, paediatrics, intensive care, etc.) in social medical organizations, during physical care for sportswomen (it may be used before or after sport, however it is not recommended to use it during exercise), etc. The product is not suitable for continuous patient control. Installing the Cells 1 Put both AAA batteries in the stack cassette with correct polarity. 2 Push the battery lid horizontally in the direction of the arrow indicated below. Notes: The batteries must be properly installed. Otherwise, damage could be caused to the aircraft. Put or remove the batteries in correct order, or it is likely to damage the attachment device. Remove the battery if Oximeter will not be used for a long time.


Use Cord 1 Insert the thinst end of cord into the notch provided for this purpose. 2 Insert the thickest end of the cord into threaded tip. Tighten. WARNING! 1 Keep pulse oxygenometer out of the reach of children. Small components such as the battery compartment cover, batteries and cord are likely to cause choking if swallowed. 2 Do not suspend the apparatus with electrical cable (cord). Accessories for Product 1 One cord 2 Two AAA 3 batteries User manual Operating instructions. 1 Insert two AAA batteries into the battery compartment respecting polarity and then close the lid. 2 Press the clamp as described on the schema. 3 Place a finger in the rubber opening of I-O-M (II is better to deeply engage your finger) before releasing the clamp. 4 Press the switch once, located on the front panel. 5 Your finger must not tremble when Ioximeter is working. We recommend you stay still. 6 Read the corresponding data on the display screen. 7 Six display modes. When you turn on the oxygen meter, each time you press the power switch, the oxometer passes into another display mode; there are six in total:

When you keep the pressure on the power switch (more than a second), the brightness of the oximeter will change by degrees; there are ten degrees of shine, with the default level being Level Four. Declaration: Please use medical alcohol to clean the rubber, located inside the oxometer, in contact with your finger and cleanse the finger for testing with alcohol before and after each test. (The rubber inside the oximeter is suitable for medical use, has no toxin and does not harm human skin). When your finger is inserted into the oximeter, the surface of your nail must be directed upwards.


Front Façade The heartbeat bar diagram corresponds to the pulse frequency. Maintenance and Storage 1 Replace batteries when the "low voltage" indicator lights up. 2 Upstream of any use as a patient diagnosis, clean the pulse oxygenmeter surface. 3 If no use is allowed for a long period, remove the batteries from the oxygen meter. 4 Preferably store the appliance between -20°C~+55°C and within an area with a moisture content ≤93%. 5 Store dry. Extreme humidity will reduce the life of l-oxymeter and even cause irreversible damage. 6 Please comply with the local and/or national provisions on scrapping used batteries. Cleaning of touch pulse l-oxymeter To clean silicone in contact with the finger and present inside the oxygen meter, it is appropriate to use a "medical" type alcohol. To do this, clean with a soft cloth impregnated at 70% isopropyl alcohol. Upstream and after each test, clean the measured finger with alcohol. Do not pour or spray liquids on the surface of the oxymeter. Avoid any penetration of liquid within the openings of the instrument. Dry the oxygen meter before re-use. A so-called "functional" tester cannot be used as a measure of the accuracy of an oximeter pulse sensor or monitor. Clinical trials were conducted to regulate the accuracy of SpO2. The measured blood saturation value of hemoglobin in transcutaneous oxygen (SpO2) is compared with the serum saturation values by arterial sampling (SaO2), determined from blood samples using a laboratory CO-oxymeter. Accuracy of sensors compared to blood samples tested by CO-oxymeter. Measurement over the normal 70~100% SpO2 range (s) expressed as a percentage. In accordance with ISO 9919 (2005), "Electrical medical devices — specific requirements for basic safety and so-called 'essential' performance of pulse oximeter equipment for medical use", accuracy should be calculated for all subjects on the basis of effective value (or ‘RMS', namely ‘root Mean Squared'). For a daily series of 15 measurements and 10 minutes per measurement, the lifespan of this device is estimated at 5 (five) years. If problems arise, stop immediately use and contact the nearest repair centre:




Manufacturer's Guidelines and Declaration – Electromagnetic Emissions – For All EQUIPMENT/SYSTEM


IEC 61000-4-8 "hospital" or "commercial" environment. Manufacturer's Guidelines and Declaration – Electromagnetic immunity – For equipment and systems not containing any essential (vital) requirements

In theory, be precisely determined. In assessing the electromagnetic environment resulting from fixed RF emitters, consideration should be given to the on-site electromagnetic study. If the field intensity measured for the position of the Puls Oxymeter (MDF300 CF3) shall be observed to verify "normal" operation. b. If an abnormal performance is observed, additional measures will be needed. Thus, we can mention the reorientation of the positioning of. Recommended separation distances between RF Mobile and Portable communication equipment and EQUIPMENT or SYSTEMS – For all Equipment AND SYSTEMS NOT COUNTING ESSENTIAL NEEDS (VITALS)

doubts about a product problem, go to the hospital in order to get an accurate diagnosis.

Cardiac does not appear · nomally