Adam Equipment MIW-20 BABY SCALE ADAM EQUIPMENT CO. LTD. p.n. 4295, Rev. B, March 2004 © Adam Equipment Company 2004.
© Adam Equipment Company 2004.
CONTENTS 1.0 INTRODUCTION ......................................................................................................... 2 2.0 TECHNICAL SPECIFICATIONS .................................................................................. 3 3.0 LOCATING THE SCALE .
1.0 INTRODUCTION • The MIW-20 scales are precision m edical scales specially designed to weigh babies on a cradle. • The scale has a capacity of 20 kg and readability of 0.01 kg. • The scale displays the weight in kilograms or pounds. • The scale is fast, accurate and simple to use.
2.0 TECHNICAL SPECIFICA TIONS Specifications MIW-20 Maximum Capacity, max 20 Kg (44.4 lbs.) Readability, d 10 g (1/2 oz.) Tare range, -T -20 kg (44.4 lbs.
3.0 LOCA TING THE SCALE • Place the scale on a stable surface. A soft and uneven floor (e.g. carpet, linoleum) could cause an erroneous weight reading. • The ambient temperature of the scale should be kept as constant as possible. Do not place the scale where it may be s ubjected to draughts or heater and direct sunlight.
5.0 KEY DESCRIPTIONS [ON / OFF/TARE] Turns the scale on or off (keep it pressed for 3 seconds). Sets a zero condition when the platform is empty. Sets the display to zero by storing the current weight in the memory to tare the scale. [UNIT] Sets the Units.
• To zero the display, press the [ON/OFF/TARE] key. • In practice when the scale is turned on, the [ON/OFF/TARE] key is pressed to zero the display, if necessary. • After the display is at zero, plac e the baby to be weighed on the cradle of the scale.
8.0 HOLD FUNCTION • Due to the baby’s movement, the we ight shown on the display will change all the time. After the baby is placed on the scale, press [HOLD] to hold the weight value. The display shows ‘ hold ’. The scale will calculate the average of the former weight values taken automatically.
10.0 CALIBRA TION PROCEDURE • Place the scale on a flat platform after inserting a PP3, 9V alkaline battery. • Press the [ON/OFF/TARE] key to turn the power ON.
Manufacturer’s Declaration of Conformity This product has been manufactured in accor dance with the harmonised European standards, following the provisions of the below stated directives: Electro Magnetic Compatib ility Directive 89/336/EEC Low Voltage Directive 73/23/EEC Adam Equipment Co.
ADAM EQUIPMENT is an ISO 9001:2000 certified global organisation with more than 30 years experience in the production and sa le of electronic weighing equipments. Products are sold through a world wide di stributor network -supported from our company locations in the UK, USA and SOUTH AFRICA.
An important point after buying a device Adams & Co MIW-20 (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Adams & Co MIW-20 yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Adams & Co MIW-20 - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Adams & Co MIW-20 you will learn all the available features of the product, as well as information on its operation. The information that you get Adams & Co MIW-20 will certainly help you make a decision on the purchase.
If you already are a holder of Adams & Co MIW-20, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Adams & Co MIW-20.
However, one of the most important roles played by the user manual is to help in solving problems with Adams & Co MIW-20. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Adams & Co MIW-20 along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center