a AUTO OFF function

O
When you switch the device on, the „AUTO OFF FUNCTION“ is activated automatically. After 20 minutes, the device switches off automatically unless one of the buttons is pressed. To activate or deactivate the device‘s „AUTO OFF FUNCTION“, press the „SET“ button while pressing and holding down the „ON/OFF“ button. For optimum measurements, please wait approximately 15 minutes after switching on the measu- ring device so that it has time to adapt to the new ambient conditions in a particular room.
If visibility is poor, switch on the display illumination. In order to pre- serve the batteries, the illumination switches off automatically after 10 seconds. To reactivate it, simply press the button again. HOLD
When you switch the device on, it shows the current ambient temperature in °C on the display. To display the value in °F, press the „Unit“ button. By pressing the „Mode“ button, you can get the device to display the highest (MAX), lowest (MIN), average (AVG) or difference value (DIF) in addition to the current measured value. As its name suggests, the „HOLD“ button enables you to hold the current measured value.
In the event of rapid temperature × uctuations (± 1°C), the sensor needs to adapt to the new condi- tions. Consequently, it takes a certain amount of time for a stable value to be established. !
Relative humidity may be deÖ ned as the ratio between the air‘s current water vapour content and the maximum that it can contain at the same temperature and constant pressure. Hot air can hold more water vapour than cold air. At 100% relative humidity, the saturation point has been reached and the air cannot hold any more water vapour.
Press the „Temp/%rH“ until a measured value in % relative humidity (%RH) appears on the display. By pressing the „Mode“ button, you can get the device to display the highest (MAX), lowest (MIN), average (AVG) or difference value (DIF) in addition to the current measured value. As its name suggests, the „HOLD“ button enables you to hold the current measured value.
In the event of rapid × uctuations in relative humidity (± 1%), the sensor needs to adapt to the new conditions. Consequently, it takes a certain amount of time for a stable value to be established. !
Activated Deactivated

A photoelectric sensor is used to measure the incident light intensity. This is the power of the radiated light per unit of area. Photometry deÖ nes light in terms of how it is perceived by the human eye. The sensitivity of the eye also depends on the wavelength or light colour. The measuring unit uses special colour Ö lters on the sensor that adapt to the sensitivity curve (CIE curve).
After switching on the device, press the „Lux“ button. Please observe the following application notices when determining illuminance: – Align the sensor on the measuring surface so it is perpendicular to
the light source – During measurement, the incidence of light must not be affected
by shadows from the operator or other objects not involved in the measurement – Remove the wind de× ector from the microphone – You can check the measuring device‘s zero position by covering the
sensor completely. – Discharge, × uorescent or energy-saving lamps should be switched
on well in advance of the measurement process (approximately 15 minutes) so that the brightness has time to stabilise
By pressing the „Mode“ button, you can get the device to display the highest (MAX), lowest (MIN), average (AVG) or difference value (DIF) in addition to the current measured value. As its name suggests, the „HOLD“ button enables you to hold the current measured value.
The device measures radiation in the main direction, perpendicular to the sensor plane. Incident light rays at an angle are taken into account by the sensor using an adjustment curve. !


Practical example Luminous intensity in lux
Sunshine 50,000 – 100,000
Overcast 2000 – 10,000
Good street lighting 20
Full moon 0.01 – 0.1
This device registers sound in a similar way to the human ear. The integrated microphone receives the incoming sound waves and converts them into electrical signals. Human hearing involves speciÖ c auditory curves. The greater the sound pressure, the louder the sound is perceived to be. The higher the frequency, the higher the pitch of the sound is perceived to be. To ensure that measurements are aurally compensated, this measuring device is Ö tted with Ö lters. The A-weighting (dB A) replicates the frequency response associated with human hearing and is used for the majority of industrial noise and environmental noise measurements. The C-weighting is used in the case of scientiÖ c measurements, for example.
After switching on the device, press the „dB“ button. To switch between dB (A) and dB (C), press the „Unit“ button. Point the microphone towards the relevant sound source and observe the relevant distance. By pressing the „Mode“ button, you can get the device to display the highest (MAX), lowest (MIN), average (AVG) or difference value (DIF) in addition to the current measured value. As its name suggests, the „HOLD“ button enables you to hold the current measured value.
Powerful wind noises (> 10m/s) can affect the measured value. In such cases, you should use the wind de× ector supplied with the product. !
Sound source Sound pressure level in dB (A)
Quiet residential area 30 to 40
Low-level conversation, quiet ofÖ ce 40 to 50
Normal conversation 50 to 60
Heavy road trafÖ c 70 to 80
Shouting, screaming 80 to 85
Pneumatic hammer (10 m away) 90 to 100
Start-up of a jet aircraft (100 m away) 120 to 130
Auditory threshold 0 · Pain threshold 140

The measuring device is capable of measuring wind speed or air × ow with the built-in fan wheel. The resulting measurements can be output in various units. On the basis of these quantities, it is also possible to measure volumetric × ow.
After switching on the device, press the „Anemo“ button. Press the „Unit“ button repeatedly to change the units: FPM (feet per minute), knots, m/s (metres per second) and km/h (kilometres per hour). Press the „Anemo“ button again if you want to determine the volumetric × ow. Use the „Unit“ button to switch between the units CMM (cubic metres per minute) and CFM (cubic feet per minute). Firstly, measure the × ow opening area. Then you can set the measured area by pres- sing the „Set“ button and making the necessary adjustments with the „Unit“ (change position of decimal point), „Hold“ (+) and „Illuminati- on“ (-) buttons. Press the „Set“ button to conÖ rm your entry.
Please observe the following application notices when deter- mining wind speed/air ƃ ow: – Avoid direct sunlight during the measurement process – Place the device as close as possible to the source of the air × ow – Position the device parallel to the source of the air × ow – Use the MAX function to determine the point where the air × ow
By pressing the „Mode“ button, you can get the device to display the highest (MAX), lowest (MIN), average (AVG) or difference value (DIF) in addition to the current measured value. As its name sug- gests, the „HOLD“ button enables you to hold the current measured value.
is greatest · (+) (–)

Functional and operational reliability are only warranted when the measuring device is operated within the speciÖ ed climatic conditions and only when used for those purposes for which it is designed. It is the user‘s responsibility to assess the measurement results and any actions taken as a consequence in accordance with the nature of the task concerned. !
Ambient temperature Resolution Accuracy
-10°C to 60°C 0.1°C ± 1.5°C
Relative humidity Resolution Accuracy Resolution Accuracy
0% to 100% r.H. 0.1% (20 to 80%) ± 3% / 25°C 0.1% (< 20% and > 80%) ± 5% / 25°C
Sound pressure level 30 to 130dB (A): Resolution Accuracy Sound pressure level 30 to 130dB (C): Resolution Accuracy Frequency range Test conditions
0.1% ± 1.5 dB 100 to 8000 Hz 94dB, 1 kHz sine curve
Illuminance At a colour temperature of 2856 K, calibrated according to reference incandescent lamp 2856K 0 to 2 klx: Resolution / Accuracy 0 to 20 klx: Resolution / Accuracy 0 to 50 klx: Resolution / Accuracy
1 lux / ± (5% of measured value + 10 digits) 10 lux / ± (5% of measured value + 10 digits) 100 lux / ± (5% of measured value + 10 digits)
Wind speed 0.5 to 20 m/s: Resolution / Accuracy 1.8 to 72 km/h: Resolution / Accuracy 1.6 to 65.7 ft/s: Resolution / Accuracy 0.9 to 38.9 knots: Resolution / Accuracy
0.1m/s / ± (3% of measured value + 10 digits) 0.1km/h / ± (3% of measured value + 10 digits) 0.1ft/s / ± (3% of measured value + 10 digits) 0.1 knots / ± (3% of measured value + 10 digits)
Air ƃ ow 0 to 999.9 CMM m/s: Accuracy 0 to 999.9 CFM ft/s: Accuracy ± (3% of measured value + 10 digits) ± (3% of measured value + 10 digits)
Power supply 1 x 9 V IEC 6F22
Operating temperature 0°C to 60°C
Dimensions (W x H x D) 280 x 89 x 50 mm
Weight 0.430 kg
Subject to technical changes without notice. 06/10
EU-Bestimmungen und Entsorgung Das Gerät erfüllt alle erforderlichen Normen für den freien Warenverkehr innerhalb der EU. Dieses Produkt ist ein Elektrogerät und muss nach der europäischen Richtlinie für Elektro- und Elektronik-Altgeräte getrennt gesammelt und entsorgt werden.
Weitere Sicherheits- und Zusatzhinweise unter: www.laserliner.com/info
Technical data