Laserliner LuxTest-Master

60 pages·870 KB·Measuring instruments·PDF·ENJA
Laserliner LuxTest-Master
Table of Contents 2 sections LuxTest-Master · Reference values for illuminance in indoor spaces ShowHide
Page 1 of the Laserliner LuxTest-Master user manual — LuxTest-Master
Page 1 · click to enlarge
Page 1

LuxTest-Master

Laserliner LuxTest-Master — LuxTest-Master diagram (page 1)
Fig. 1 · LuxTest-Master

Read the operating instructions and the enclosed brochure „Guarantee and additional notices“ completely. Follow the instructions they contain. Safely keep these documents for future reference. !

Function/application The measuring instrument is used to measure illuminance in workplaces, ofÖ ces, public buildings and in industry. The integrated photodiode determines the illuminance in lux and foot-candles. With the corresponding function, it is possible to carry out comparison and reference measurements as well as peak measurements for registering quick changes in light levels.

Laserliner LuxTest-Master — LuxTest-Master diagram (page 1)
Fig. 2 · LuxTest-Master
2Relative measurement 3 4 5MAX/MIN function PEAK function Low battery charge 6Analogue scale 7Units: Lux, foot-candle 8Measured value 9 10Automatic measuring range Hold measured value
Laserliner LuxTest-Master — LuxTest-Master diagram (page 1)
Fig. 3 · LuxTest-Master
3Hold measured value / zeroing function 4Relative / peak measurement 5Unit of measure 6MAX/MIN function 7ON/OFF / button tones 8LC display
Laserliner LuxTest-Master — LuxTest-Master diagram (page 1)
Fig. 4 · LuxTest-Master
Page 2 of the Laserliner LuxTest-Master user manual — LuxTest-Master
Page 2 · click to enlarge
Page 2
Laserliner LuxTest-Master — LuxTest-Master diagram (page 2)
Fig. 5 · LuxTest-Master

Inserting the batteries Open battery compartment and insert batteries correspon- ding to installation symbols. Ensure correct polarity.

Laserliner LuxTest-Master — LuxTest-Master diagram (page 2)
Fig. 6 · LuxTest-Master

With the device switched on, briefly press and hold the „ON/OFF“ button to deacti- vate the button tones. Press the button again to reacti- vate the button tones.

After being switched on, the measuring instru- ment always starts set to the automatic measu- ring range. Press the „MAN/AUTO“ button to change over to manual measuring range. The measuring range, i.e. the decimal place (20.00 lux; 200.0 lux; 2,000 lux; 20,000 lux; 200,000 lux) changes with each additional press of the button. „OL“ will appear in the display when the measured value is outside the measuring range. Press the „MAN/AUTO“ button until the corresponding manual measuring range is set or until „AUTO“ appears in the display, indicating that the automatic measuring range is set again.

Page 3 of the Laserliner LuxTest-Master user manual — LuxTest-Master
Page 3 · click to enlarge
Page 3

LuxTest-Master

Laserliner LuxTest-Master — LuxTest-Master diagram (page 3)
Fig. 7 · LuxTest-Master

After switching on the device, activate relative measurement by pressing the „REL/PEAK“ button. The value last displayed will be set as the reference value and it now shows the difference with respect to the set reference value. The relative measurement can be carried out both in the automatic as well as in the manual measuring range. This function is deactivated by pressing the „REL/PEAK“ button again.

Laserliner LuxTest-Master — LuxTest-Master diagram (page 3)
Fig. 8 · LuxTest-Master

After switching on the device, activate peak measurement by pressing the „REL/PEAK“ button. This function increases the response of the sensor, thus enabling it to register quick changes in light levels. Peak measurement is only possible in the manual measuring range. Set the corresponding measuring range as described in Step 4. Press and hold the „REL/PEAK“ button to deactivate this function.

Laserliner LuxTest-Master — LuxTest-Master diagram (page 3)
Fig. 9 · LuxTest-Master

The peak value remains visible on the display. Measured values lower than the peak value will not be displayed. !

5 The measured values can be displayed in Lux and FC (foot-candles). Press the „UNIT“ button to change the unit of measure.

Page 4 of the Laserliner LuxTest-Master user manual — Reference values for illuminance in indoor spaces
Page 4 · click to enlarge
Page 4
Laserliner LuxTest-Master — Reference values for illuminance in indoor spaces diagram (page 4)
Fig. 10 · Reference values for illuminance in indoor spaces

The zero point of the sensor curve can be adjusted for optimum measurement results at low illuminance levels. After switching on the device, Ö t the protective cap in order to completely cover the sensor. Press and hold the „HOLD/ZERO“ button to start the adjust- ment procedure. Wait until an acoustic signals sounds from the device to conÖ rm successful adjustment. The value 0.00 lux appears on the display. Avoid strong sunlight while adjusting the device.

MAX/MIN The MAX/MIN function is used to show the highest and lowest measured value during a measurement. To deactivate the function, press and hold the „MAX/MIN“ button until the required function appears on the display or until „MAX“ or „MIN“ is no longer visible.

Laserliner LuxTest-Master — Reference values for illuminance in indoor spaces diagram (page 4)
Fig. 11 · Reference values for illuminance in indoor spaces

Reference values for illuminance in indoor spaces

Laserliner LuxTest-Master — Reference values for illuminance in indoor spaces diagram (page 4)
Fig. 12 · Reference values for illuminance in indoor spaces
actical exampleLuminous intensity in lux afÖ c areas, ancillary rooms20 orridors in buildings used by people50 nitary facilities, machine rooms, staircases100 gularly occupied work stations within production cilities, canteens200 onference rooms, gyms300 Ö ces, medical facilities (emergency unit)500 sign and drawing rooms750 onitoring locations, assembly rooms, test stations1000 sembly rooms for small components1500

0 2 sec

Page 5 of the Laserliner LuxTest-Master user manual — LuxTest-Master
Page 5 · click to enlarge
Page 5

LuxTest-Master

Technical data Measuring range20.00 lux; 200.0 lux; 2,000 lux; 20,000 lux; 200,000 lux Accuracy3% V(λ) adaptation 2% cosine correction SensorSilicon photodiode Spectral range320 nm…730 nm LC display3 1/2 positions with analogue bargraph Sampling rateø 2 measurements/second StandardDIN 5032-7 Type B Voltage supply1x 6F22 9V Operating temperature-10 °C ... 50 °C Max. relative humidity85% (no condensation) Dimensions (W x H x D)89 mm x 190 mm x 42.5 mm Weight250 g (including batteries)

Technical revisions reserved. 02.12

EU directives and disposal This device complies with all necessary standards for the free movement of goods within the EU. This product is an electric device and must be collected separately for disposal according to the European Directive on waste electrical and elec- tronic equipment. Further safety and supplementary notices at: www.laserliner.com/info

Ask about the Laserliner LuxTest-Master

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the Laserliner LuxTest-Master

Be the first to ask — other owners or our team will answer.

Original document

Laserliner LuxTest-Master manual PDF

60 pages·870 KB·Measuring instruments·PDF
Click any page above to open it here
Page 1 / …
100%
Loading pages…