Laica TH2002

26 pages·741 KB·Thermometer·PDF·ENNLJA
Laica TH2002
Table of Contents 3 sections GB English EAR INFRARED THERMOMETER GB English Instructions and guarantee · INSTRUCTIONS FOR CORRECT USE · TECHNICAL DATA ShowHide
Page 1 of the Laica TH2002 user manual — GB English EAR INFRARED THERMOMETER GB English Instructions and guarantee
Page 1 · click to enlarge
Page 1

GB English EAR INFRARED THERMOMETER GB English Instructions and guarantee

Before using it, it is important to read carefully in all their parts the instructions and warnings contained in this booklet. Preserve the booklet with care. Your new ear thermometer detects body temperature in an absolute precise and painless way. It is ideal for taking babies' temperature, even when they sleep. The new measurement infrared technology results into greater reliability and above all longer instrument life duration. Thanks to the wide LCD-display, it allows an easy reading. It is extremely easy to use and equipped with acoustic signal when body temperature measurement has been completed. It stores the last measurement performed. Laica S.p.A. - Viale del Lavoro, 10 - Fraz. Ponte 36021 Barbarano Vicentino (VI) - ITALY Fax: +39-0444-795324 Made in China

3) Keep out of reach of children. 4) Do not expose the thermometer to extreme temperatures, direct sunlight and do not store it close to heat sources. Store in a dry and clean place. 5) Do not touch the probe with the fingers and do not disassemble the thermometer. 6) Do not immerse the device in water or other liquids. 7) Do not use the ear thermometer to take rectal, oral or axillary temperature. 8) Measurement accuracy may be jeopardized by: taking vasoconstrictive medicines and ear irritations. 9) Before each measurement, clean the ear where the probe will be positioned. 10) Take the temperature away from direct sunlight and wind. 11) This thermometer has to be used to measure ear temperature as shown in figure 1. 12) Keeping the thermometer in hand for too long can cause the probe to measure a lower body temperature. 13) Do not take the temperature for thirty minutes after sports activity or having had a bath.

  • In fact, it represents a defensive reaction of our organism that is characterised by an increase in body temperature above 37°C and by an increase in heartbeat and breathing frequency. The normal temperature inside our organism is around 37.5°C, which is the value read with the rectal method, the temperature in the mouth is lower by about 0.5°C (37°C), while that under the armpit is lower by 1°C (36.5°C). Therefore it is very important that the temperature is correctly measured using the most effective method, and preferably not in those moments when body temperature could be higher (on waking, after meals). The temperature of a healthy person is also affected by the following factors:
  • The person’s individual level (individual metabolism).
  • Age (body temperature is higher in babies and small children and falls with increasing age. Greater temperature fluctuations occur in children faster and more often.).
  • Clothing.
  • The outside temperature.
  • The time of day (Body temperature is lower in the morning and rises as the day goes on).
  • Preceding physical activity.
  • Measuring method.
  • Phase of the menstrual cycle. It is necessary to consider that skin temperature can be affected particularly by the room temperature. Underarm temperature measurement is, in fact, less reliable than oral and rectal measurement.

The temperature measured in the ear is reliable because the thermometer detects the heat produced by the eardrum. It detects the temperature of the nervous centre located in the brain, the area regulating body temperature. This measurement method is very appreciated as it is non-invasive, accurate and reliable. Please remember, that the thermometer must be in the room where the temperatures is to be taken at least half an hour beforehand, while the person must be present at least 10 minutes beforehand. The table below is a eist of querage temperatures "normal". However, we recommend that you learn to recognise your normal body temperature when you are well, to correctly interpret the different readings. MEASURING METHOD NORMAL AVERAGE ARMPIT 36.5°C ORAL 37°C RECTAL 37.5°C EAR 37°C

GENERAL WARNINGS 1) Self-measurement means control, not diagnosis or treatment. Unusual values have always to be discussed with one's own physician. Do not modify, in any case, dosages of any drug prescribed by your doctor. 2) This thermometer is a sensitive instrument. Handle it with care and do not expose it to mechanical bumps. Do not press, bend, let fall or break into pieces the thermometer.

These different thermometers are suitable for measuring the temperature in different parts of the body: forehead thermometer: measurement only on the forehead, ear thermometer: measurement only in the ear, rod thermometer: rectal (conventionally in the back passage), axillary (under the arm) or buccal measurement (inside the cheek) and under the tongue (sublingual). The

A HIGH TEMPERATURE A high temperature should be considered a symptom and not an illness. Usually it is the sign that our organism is fighting against an infection.

Page 2 of the Laica TH2002 user manual — INSTRUCTIONS FOR CORRECT USE
Page 2 · click to enlarge
Page 2

GB English EAR INFRARED THERMOMETER GB English Instructions and guarantee

temperature varies depending on the part of the body where it is measured. The variation can be between 0.2 – 1° C. Temperatures measured with different thermometers cannot be compared with one another. Tell your doctor or take into account when diagnosing yourself what thermometer you used in what part of the body to measure the temperature.

5. Then, remove the probe from the ear; read the measured value on the display. When the detected temperature is lower than 38.0°C, the display shows the drawing “ ” (see fig.5); while when it exceeds this value “ ” appears (see fig.6). Note: The temperature shown on the display corresponds to oral temperature. 6) The device turns off automatically after one minute if it is not used for such time.

Read the paragraph "Taking ear temperature" again. The probe could be dirty, clean it before and/or after each use with a cotton swab soaked in alcohol.

INSTRUCTIONS FOR CORRECT USE

Drawing 1 Problem: Flat battery. Solution: Replace the battery.

MEASUREMENT OF EAR TEMPERATURE Caution! In order to assure an accurate temperature measurement, clean the any part of the probe using a cotton swab soaked in alcohol after each use. Make sure that the probe is perfectly dry before use. AN IMPROPER CLEANING OF THE PROBE CAN RESULT IN NOT RELIABLE TEMPERATURE MEASUREMENTS. 1) Clean the ear where the probe will be positioned. 2) Switch on the thermometer by pressing the key O/I; the device emits a short acoustic signal. On the display all function symbols do switch on, just for few seconds. Then, the thermometer displays for few seconds the last measurement performed (see fig. 2). 3) To obtain an accurate measurement, gently pull the ear towards outside in order to straighten the ear canal (see fig. 3) and let the probe adhere to this latter pointing at the eardrum membrane (see fig. 4). 4) Keep the "O/I" key pressed until a long acoustic signal indicates that the measurement has ended (the time necessary to perform a measurement ranges between 5 and 30 seconds according to the person's temperature).

INSERTING/REPLACING THE BATTERY The battery must be replaced urgently when the display shows the symbol of “ ” (see drawing 1). Before replacing the battery, check that the thermometer is off to avoid damaging the mechanism. 1) Let the cover of the battery compartment slide. 2) Insert a pointed object in the battery compartment, such as a pen, and lift the battery trying to remove it. 3) Put a new battery under the metallic hook on the left and press downwards the battery towards the right until a click is heard. The positive pole must be positioned upwards. 4) Reassemble the cover on the battery compartment. Remove run down batteries by sliding the cover of the batteries compartment, following the arrow direction. Extract the batteries and dispose them as special waste at the specific collection points forecast for recycling. For more information about the disposal of run down batteries, please contact either the store where the device containing the batteries was purchased, or the Municipality, or the waste disposal local service.

CLEANING AND MAINTENANCE Before and/or after each use clean the thermometer probe with a cotton swab soaked in alcohol to assure reliable measurements and to avoid microbial contaminations. The probe is the most delicate part of the thermometer. Pay utmost care to it during the cleaning. Let the probe dry completely for at least 30 minutes. The thermometer body has to be cleaned using a soft and dry cloth. DO NOT WASH THE THERMOMETER BODY WITH WATER. Absolutely do not use abrasive products and do not immerse the thermometer in water or other liquids. Store the thermometer in a dry and clean place, away from direct sunlight.

Drawing 2 Problem: The display shows "Lo" or "Hi". Room temperature is not included between +10°C and +50°C. Solution: It is necessary to consider that the thermometer has to be already in the room, in which the measurement will be carried out, half an hour before the actual measurement performance; while the person should be in the room for at least 30 minutes before performing the measurement. Room temperature has to range between +10°C and +50°C.

Drawing 3 Problem: The display shows “Err”. Room temperature is not included between +16°C and +40°C. Solution: It is necessary to consider that the thermometer has to be already in the room, in which the measurement will be carried out, half an hour before the actual measurement performance;

POSSIBLE PROBLEMS AND RELATED SOLUTIONS Problem: The thermometer detects temperature values lower than the

actual ones Solution: Check the correct positioning of the probe in the ear canal.

Page 3 of the Laica TH2002 user manual — TECHNICAL DATA
Page 3 · click to enlarge
Page 3

GB English EAR INFRARED THERMOMETER GB English Instructions and guarantee

Laica TH2002 — GB English EAR INFRARED THERMOMETER GB English Instructions and guarantee diagram (page 3)
Fig. 1 · GB English EAR INFRARED THERMOMETER GB English Instr…

while the person should be in the room for at least 30 minutes before performing the measurement. Room temperature has to range between +16°C and +40°C.

Laica TH2002 — GB English EAR INFRARED THERMOMETER GB English Instructions and guarantee diagram (page 3)
Fig. 2 · GB English EAR INFRARED THERMOMETER GB English Instr…

ELECTROMAGNETIC COMPATIBILITY This device complies with current regulations relating to electromagnetic compatibility (EMC) and has been designed to be used in the home. Its emissions are extremely small and are unlikely to generate interference with other equipment. In the event that it is to be used in close proximity to other equipment, it is recommended to follow the instructions given in the tables at the end of the instructions for use. Certain mobile communications devices and radio systems may have an effect on the operation of this device. In cases where there are anomalies during operation, it is recommended to distance the device from any other equipment that could cause electromagnetic interference and check for the restoration of performance. In all cases, in the event of doubt, contact Customer Support or see the tables relating to electromagnetic emissions at the end of these instructions.

a retail location that is over 400 m² without having to purchase a new, similar device. This procedure of separated collection of electric and electronic devices is carried out forecasting a European environmental policy aiming at safeguarding, protecting and improving environment quality, as well as avoiding potential effects on human health due to the presence of hazardous substances in such equipment or to an improper use of the same or of parts of the same. Caution! The wrong disposal of electric and electronic equipment may involve sanctions.

does not apply to parts subject to tear and wear along the use, as well as to batteries, when they are included. Warranty decays after 2 years from the purchase date. In this case, technical service interventions will be carried out against payment. Information on technical service interventions, under warranty terms or against payment, may be requested by writing at info@ laica.com. Repair or replacement of products covered by the warranty terms are free of charge. In case of faults, refer to the dealer; DO NOT send anything directly to LAICA. All interventions covered by warranty (including product replacement or the replacement of any of its parts) shall not be meant to extend the duration of the original warranty period of the replaced product. The manufacturer cannot be made liable for any damages that may be caused, directly or indirectly, to people, things and pets resulting from the inobservance of the provisions contained in the suitable user manual, especially with reference to the instructions related to installation, use and maintenance of the equipment. Being constantly involved in improving its own products, Laica reserves the right to modify its product, totally or partially, with reference to any production need, without that this involves any responsibility whatsoever by Laica or its dealers.

TECHNICAL DATA

IP22Degree of protection of electrical equipment covers, where the first

figure indicates the degree of protection against penetration by solid foreign bodies (from 0 to 6) and the second figure the degree of protection against penetration by liquids (from 0 to 8).

WARRANTY This equipment is covered by a 2-year warranty starting from the purchase date, which has to be proven by means of the stamp or signature of the dealer and by the fiscal payment receipt that shall be kept enclosed to this document. This warranty period complies with the laws in force and applies only in case the consumer is an individual. Laica products are designed for domestic use only; therefore, their use for commercial purposes is forbidden. The warranty covers only production defects and does not apply, if the damage is caused by accidental events, misuse, negligence or improper use of the product. Use the supplied accessories only; the use of other accessories invalidate the warranty cover. Do not open the equipment for any reason whatsoever; the opening or the tampering of the equipment, definitely invalidate the warranty. Warranty

Battery: 1 x CR2032LI Display: liquid crystals Switching-off: automatic, after one minute, if not used. Measurement range: 10°C to 50°C Resolution: 0.1°C Operation conditions: 16°C to 40°C Storage conditions: 10°C to 50°C Tolerance for body temperature: from 36°C to 39°C; +/-0.2°C 39°C; +/- 0.3°C

DISPOSAL PROCEDURE (Dir. 2012/19/Eu-WEEE)

The symbol on the bottom of the device indicates the separated collection of electric and electronic equipment. At the end of life of the device, do not dispose it as mixed solid municipal waste, but dispose it referring to a specific collection centre located in your area or returning it to the distributor, when buying a new device of the same type to be used with the same functions. If the appliance to be disposed of is less than 25 cm, it can be returned to

Sources

References & links

Pages referenced in this manual — manufacturer, support and downloads.

Ask about the Laica TH2002

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the Laica TH2002

Be the first to ask — other owners or our team will answer.

Original document

Laica TH2002 manual PDF

26 pages·741 KB·Thermometer·PDF
Click any page above to open it here
Page 1 / …
100%
Loading pages…