Muon Liquid Handling System User Guide C. Johnson, S.P. Cottrell et al Version 0.2 - 1 -.
Contents 1 Getting Started 3 1.1 Layout of the liquid handling system 3 1.2 Layout of the in situ sample stick 4 1.3 Layout of the pump 4 2 General procedures 6 2.1 Sample loading 6 2.1.1 Loading Vessel 1 6 2.1.2 Loading Vessel 2 6 2.2 Evacuation of the system 6 2.
1 Getting S t arted This manual describes the liquid handling system and the in situ sample cell as use d on the DEVA instrument with the “RF” spect rometer, for either norm al muon spins relaxation or RF resonance experiments.
1.2 Layout of the in situ sample stick The liquid sample stick is designed to fit into the DEVA f low cryostat, details of which can be found in the DEVA manual. It consists of a shapol target cell 30 mm x 30 mm with a mylar window upon which may be mounted an RF coil.
Roughing/ Turbo switch Off/Run switch Pressure reading Turbo control panel Turbo Pump Rotary Pump Figure 2 The lay out of the fr ont panel o n the vacuum pump use d with the muon liqui d handlin g system.
2 General procedures 2.1 Sample loading At this point, the system will be open to the atm osphere, ensure that valve V2 is closed to prevent air cond ensation in the cold trap. 2.1.1 Loading Vessel 1 • Open the tap on the right-hand side of Vessel 1.
2.2.1 Removing air from Vessel 1 Once the system has been pressuris ed, tap A on Vessel 1 can be opened. It is possible to bubbled helium gas through the sample liqui d against the non-retu rn valve, V13. If this is required, first ensure th at valves V 5, V7 and V8 are closed.
2.3 Freeze-Pump-Thaw cycle Using this system it is possible to o perat e freeze-pump-thaw cycles on sam ple liquids held in any of the three vessels. Procedur es appropriate to ea ch vessel are outlined below. 2.3.1 Sample in Vessel 1 • All taps and valves should be closed.
2.3.3 Sample in Vessel 3 • All taps and valves should be closed. • Open valves V2 and V3 along with tap C on Vessel 3 and freeze the sample liquid by slowly raising a dewa r of liquid nitrogen around it. • When it has been determined that the liquid ha s completely frozen switch the pump to roughing mode and open valve V1.
2.4.2 Vessel 3 to sample cell The transfer of liquid from Vessel 3 to the sample cell is brought about by the difference in pressure between that in the Vessel and a vacuum in the cell. However, in order to control the transfer of liquid th e pressure difference shou ld be as small as possible but not so low that the liquid starts to boil.
2.4.4 Recovery of liquid in sample cell to Vessel 1 The liquid in the sample cell cannot be recovered to Vessel 1 using the liquid handling rig as presently configured. 2.4.5 Recovery of liquid in sample cell to Vessel 2 The liquid in the sample cell can be r ecovered to Vessel 2 using the following procedure.
• All the liquid may not return in a single pass and repea ting the evacuation procedure may be required. However, some liquid m ay be present in the system and care should be taken during evacuation that th e trap does not become blocked. The liquid may also be encouraged to leave the sample cell by using a small amount of gas.
3 Example Experiment The paramagnetic signal in liquid n-hexane is difficult to observe because the muonium polarisation is found to decay on a microsecond time scale, which is further reduced by the presence of dissolved oxygen.
0.0 0.5 1.0 1.5 2.0 2.5 3.0 11.0 11.5 12.0 12.5 13.0 13.5 Initial asymmetry (%) Time ( µ s) 0.0 0.5 1.0 1.5 2.0 2.5 3.0 11.0 11.5 12.0 12.5 13.0 Initial asymmetry (%) Time ( µ s) 1.
An important point after buying a device Boca Research None (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Boca Research None yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Boca Research None - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Boca Research None you will learn all the available features of the product, as well as information on its operation. The information that you get Boca Research None will certainly help you make a decision on the purchase.
If you already are a holder of Boca Research None, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Boca Research None.
However, one of the most important roles played by the user manual is to help in solving problems with Boca Research None. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Boca Research None along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center