Geratherm Micromesh

164 pages·13.3 MB·Inhaler·PDF·EN
Geratherm Micromesh
Table of Contents 23 sections Items Supplied and Packaging · Features · Descriptions of Parts · Safety Precautions · … ShowHide
Page 1 of the Geratherm Micromesh user manual
Page 1 · click to enlarge
Page 1

Thank you for your confi dence in Geratherm and congratu-

lations on your purchase! By purchasing the Geratherm micromesh nebulizer you have acquired a quality product by Geratherm®. Please read these instructions for use carefully to use the device without problems and enjoy your Geratherm® neb- ulizer for a long time.

Best Regards, The Geratherm® Team

Dear customer,

Page 2 of the Geratherm Micromesh user manual
Page 2 · click to enlarge
Page 2
Page 2 of the Geratherm Micromesh user manual
Page 3 of the Geratherm Micromesh user manual — Features
Page 3 · click to enlarge
Page 3

Items Supplied and Packaging

Before you start to use the Geratherm® micromesh, please check if the following items are included in the package:

  • 1 nebulizer Geratherm ® micromesh GT-7002 with

nebulizer unit, mesh cap, adapter (for masks and

mouth piece) and cover

  • 1 mouthpiece*
  • 1 PVC mask adult*
  • 1 PVC mask children*
  • 2 “AA” batteries 1.5 V
  • 1 storage bag
  • 1 instruction manual

* Note For reasons of hygiene, please use the enclosed masks and mouth piece for one specifi c in- dividual only and do not exchange them. The masks and the mouth piece must be cleaned after each use as described in the section entitled “Cleaning and Disinfection of Nebuliser Parts/ Accessories“. Only silicon masks, which are available from specialist suppliers, can be disinfected by boiling or by means of vaporisation (at 100 °C for approx. 10 minutes).

Use only the parts listed above; use of other parts may con- stitute a hazard.

Features

  • Discharge-proof - fully operative even when positioned

on an inclined surface (up to 45°) ■ No warming of medicine, thus preventing damage to the

  • structure of the medication
  • Low power consumption
  • Convenient design – compact and lightweight, ideal for
  • travelling
  • Silent operation – thanks to the micromesh technology
  • Short treatment time – with high nebulization rate
  • No medication cups necessary (no disposable items)
Page 4 of the Geratherm Micromesh user manual — Descriptions of Parts
Page 4 · click to enlarge
Page 4

Descriptions of Parts

Geratherm Micromesh — Descriptions of Parts diagram (page 4)
Fig. 1 · Descriptions of Parts

Indicator: green: usage red: battery needs to be replaced green and red: reset the vibration rate

Page 5 of the Geratherm Micromesh user manual — Safety Precautions
Page 5 · click to enlarge
Page 5

Safety Precautions

  • Use the equipment only for its intended use as described in this manual. Do not use attachments not recommended by the manufacturer.
  • Use the equipment with medications only under the instruction of your physician.
  • Do not use the equipment if it has any damaged parts, or it has fallen into water/ other liquids.
  • Remove the batteries when not in use for a long period of time to avoid circuit damage due to leakage from battery.
  • To use the equipment for the fi rst time, or after storing it for an extended period, be sure to clean and sterilize all necessary parts as described in the cleaning instructions.
  • Do not operate the device in the vicinity of high-frequency electromagnetic transmitters.
  • Do not use while bathing.
  • The unit should not be left unattended while plugged in or turned on.
  • Keep the unit out of reach of small unsupervised children. The small parts detached from the device may result chil-
  • dren choking from inhaling or swallowing.
  • The accessible materials used in the device will not cause the potential allergic reactions to skin.
  • Do not try to modify the device to prevent any dangers.
  • Do not place the device in liquid, nor put it where it could fall into liquid. If the device becomes wet, unplug the device before touching it.
  • Do not use the device if it is not working properly, or if it has suffered any damage.
  • Do not subject the nebulizer or the mesh cap to any impacts and do not drop them.
  • Before using the nebulizer for the fi rst time, after a period of no use, and after every use, please clean all removable parts of the nebulizer.
  • Before storing, please make sure all parts are clean and dry.
  • Do not expose the nebulizer to direct sunlight, high temperatures or high humidity.
  • The device is intended for private domestic use only.
  • To guarantee an effective inhalation, just inhale with mask or mouth piece.
  • A mask and / or a mouth piece can only be used by one person.
Page 6 of the Geratherm Micromesh user manual — Important Information about Inhalants/ Medications
Page 6 · click to enlarge
Page 6

Important Information about Inhalation Therapy and Mesh Technology, Purpose of Use

Geratherm® micromesh delivers the inhalant suitable for the treatment of asthma, chronic bronchitis and other respi- ratory conditions affecting the upper and lower airways, e.g. the common cold. The medication is nebulized by means of an ultra-modern

vibrating membrane or mesh. This membrane is partially porous, enabling tiny droplets

of the inhalant to penetrate it as soon as it begins vibrating. The key component of this system is the mesh cap.

We recommend replacing the cap approx. after six months when in regular use in order to ensure effi cient nebulization.

The device is ideally suitable for inhalation at home or when

travelling. It guarantees highly effective, fast-acting inhala- tion treatment for patients of all ages. Carry out inhalation in a quiet and relaxed state, and inhale slowly and deeply, so that the medication can penetrate to the fi ne, deep bronchial tubes. Exhale normally.

Important Information about Inhalants/ Medications

Please only use the medicinal products/medication pre- scribed for you by your doctor.

The Geratherm® micromesh nebulizer is suitable for use

with all common aerosol products for medical use with the exception of

- viscous solutions (high viscosity) - suspensions (products with solid constituents) - volatile oils

Page 7 of the Geratherm Micromesh user manual — Operating Instructions
Page 7 · click to enlarge
Page 7

Preparation before Use / Operation

The nebulizer is powered by two 1.5V AA alkaline batteries.

Before the fi rst use of Geratherm micromesh the vibration speed of the membrane must be set correctly. To do this, follow the steps described under “Maintenance” on page 36.

Battery Replacement Replace batteries when the red indicator blinks. Before re- placing the batteries, make sure that the nebulizer is turned off.

1. Press the buckle on the battery cover and lift up to remove the cover. 2. Remove the used batteries. Insert two new ones. Correctly align the polarities (+ and -) with battery indication marks on the device. 3. Close the battery cover by pressing down fi rmly until a click is heard.

Keep batteries away from small children. Do not throw the batteries into fi re, they may explode.

Operating Instructions

  • Ensure that all parts are clean or disinfect if necessary and dry.
  • Before starting treatment, talk to your physician about the duration, dosage and frequency of use of the nebulizer.
  • The green indicator is on during nebulization.
  • The device turns off automatically after 15 minutes of non-use (without the ON/OFF button being pressed).
  • Do not remove the medication cup cover from the unit during nebulization.
Page 8 of the Geratherm Micromesh user manual — Operating Instructions
Page 8 · click to enlarge
Page 8

Operating Instructions

Unit Assembly and Usage

1. Remove the unit lid and the mesh cap of the nebulizer. Remove the mesh cap by gently pressing on the protruding plastic fl ap. The cover can then be pulled upwards and lifted off.
2. Slowly pour the prescribed medication into the medication cup. The maximum capacity of
Geratherm Micromesh — Operating Instructions diagram (page 8)
Fig. 2 · Operating Instructions

the medication cup is 6 ml. Please note that the medi- cation must not exceed this maximum level marking (6 ml).

3. Slide the mesh cap into the unit, align the 2 knobs with the cutouts in the direction indicated by arrow 1. Then press the cover down until it clicks into place.
4. Attach the adapter on the mesh cap. Please pay attention that the knobs of mesh cap are in the same position as the cuttings of adapter. Click the adapter into mesh cap. Now attach the unit lid.
Page 9 of the Geratherm Micromesh user manual — Operating Instructions
Page 9 · click to enlarge
Page 9

Operating Instructions

5. Attach the mask or the mouthpiece to the adapter:
Geratherm Micromesh — Operating Instructions diagram (page 9)
Fig. 3 · Operating Instructions
6. Take the nebulizer in your hand and press the ON/OFF button to begin nebulization. You can select one of two delivery modes. Please read the section “Delivery mode“ concerning this.
7. To end the nebulization, press the ON/OFF button to turn it off.

The nebulizer provides you with two delivery modes accord-

ing to your needs:

automatic or manual.

Automatic delivery of aerosol 1. Press and release the ON/OFF button. The unit turns on and starts to deliver the prescribed medication continu- ously for 15 minutes. The nebulizer will turn off automati- cally after 15 minutes of non-use (without the ON/OFF button being pressed). 2. To stop the treatment at any time, press the ON/OFF but- ton again to turn off the device.

Delivery mode

Page 10 of the Geratherm Micromesh user manual — Operating Instructions
Page 10 · click to enlarge
Page 10

Operating Instructions

Manual delivery of aerosol In the manual mode, the duration of nebulization is control- led by holding down the ON/OFF button.

1. Press the ON / OFF button for more than 2 seconds and hold. The unit starts with the nebulization, the inhalant is only set free if the on / off button is pressed. 2. When you want to fi nish the process, release the ON/ OFF button and the device will stop delivering medications.

Inhalation with Mouth Piece and Masks

Inhalation using the mouthpiece

1. Place the mouthpiece in your mouth with the lips sealed around the mouthpiece. 2. Switch on the unit. 3. During treatment, take slow, deep breaths through your mouth. If possible, hold each breath for two to three sec- onds before breathing out. This allows the Inhalant to settle into the respiratory system. 4. Continue the treatment until the Inhalant has been used up (about 10 to 15 minutes). 5. If you become dizzy, stop the treatment and rest for about fi ve minutes. Then continue the treatment, but try to breathe more slowly. If these symptoms continue with future treatments, inform your physician. 6. Clean and disinfect if necessary like described below.

Inhalation Using the Mask

1. Place the mask over your mouth and nose. 2. Switch on the unit. 3. During treatment, take slow, deep breaths through your nose. If possible, hold each breath for two to three seconds before breathing out. This allows the Inhalant to settle into the respiratory system. 4. Follow the procedures described under “Inhalation using the mouthpiece”.
Page 11 of the Geratherm Micromesh user manual — Cleaning and Disinfection
Page 11 · click to enlarge
Page 11

Cleaning and Disinfection

Please clean your device after each use according to the instructions below.

Before Cleaning Make sure that the nebulizer is turned off before disas- sembly.

1. Remove the unit lid, mask, mouthpiece and adapter from the unit. You could remove the adapter from mesh cap by slightly turning in clockwise direction.
Geratherm Micromesh — Cleaning and Disinfection diagram (page 11)
Fig. 4 · Cleaning and Disinfection
2. Remove the mesh cap by gently pressing the fl ap and lifting up in the direction indicated the arrows.
3. Discard any remaining inhalant (pay attention to statutory provisions).

Cleaning and Disinfection of Nebulizer Parts/ Accessories

Cleaning after every use Clean the mask, mouthpiece, unit lid, adapter and mesh cap after each use.

1. Disassemble the components from the unit.
2. Rinse all above-named parts under a small stream of tap water.
Geratherm Micromesh — Cleaning and Disinfection diagram (page 11)
Fig. 5 · Cleaning and Disinfection
3. Gently wipe off excess water with a clean dry cloth and allow to air dry. Possible remained drops on the mesh surface could remove by easy shaking of mesh cap.
Page 12 of the Geratherm Micromesh user manual — Cleaning and Disinfection
Page 12 · click to enlarge
Page 12

Cleaning and Disinfection

Disinfection by boiling: 1. Silicon masks and mesh cap can be disinfected by boil- ing in distilled water. 2. After 10 - 20 minutes of boiling, gently wipe off excess water with a clean dry cloth and allow to air dry on a clean paper towel.

Disinfection by using alcohol: 1. PVC mask, mouthpiece, unit lid and mesh adapter can be disinfected by using medical alcohol (Ethanol 75% dilution with water). 2. Disinfect the above-named parts by wiping with a cloth moistened with medical alcohol (Ethanol 75% dilution with water). 3. Let them air dry on a clean paper towel.

Cleaning and Disinfection the Nebulizer Unit

Cleaning after every use 1. Use a cloth moistened with water to clean the medication cup and outer casing of the unit. 2. Wipe off the unit with an unused dry cloth and let it air dry.

Daily disinfecting 1. Use a cloth moistened with medical alcohol (Ethanol 75 % dilution with water) to disinfect the medication cup and outer casing of the unit. 2. Let them air dry on a clean paper towel.

Daily disinfecting

Page 13 of the Geratherm Micromesh user manual — Maintenance
Page 13 · click to enlarge
Page 13

Cleaning and Disinfection

not hold it under running water. Be careful to ensure that no water penetrates the nebulizer unit.

  • Do not clean the nebulizer or its component parts in a

dishwasher.

  • Do not allow any moisture to come into the battery com-

partment.

  • Do not use any other disinfectant like described above to

clean the unit or its components.

  • Do not use a hairdryer to dry any parts.
  • Do not clean the unit and the accessories with brine

(NaCL).

Maintenance

When using a new mesh cap for the fi rst time, and after changing the batteries you need to reset the vibration rate:

1. Pour water into the medicine cup (pay attention to maximum marking).
2. Attach the mesh cap to the nebulizer unit.
3. Take the batteries out.
4. While holding down the ON/OFF button, reinsert the batteries and close the battery cover.
5. Keep the unit upright and then release the ON/OFF button. The green and red indicator light up, indicating that the nebulizer is resetting the vibration rate.
6. After a few seconds, the nebulizer automatically turns off. The vibration rate resetting is fi nished. The nebulizer is

ready for use.

The unit must be kept upright during this procedure.

Page 14 of the Geratherm Micromesh user manual — Changing Accessories
Page 14 · click to enlarge
Page 14

Changing Accessories

We recommend that the mesh cap be replaced after about 6 months of use if the unit is used daily (once per day), so as to ensure effi cient nebulization. This recommendation is based on a content volume of 2 ml and an inhalation time of about 8-15 minutes. The possible use time of 6 months could shorten if the unit

is used often daily.

The accessories should be replaced after continuous use.

Accessories of the nebulizer micromesh involve: mesh cap, mesh adapter (for masks and mouth piece), PVC masks (for adult and children), silicon masks (for adult and chil- dren), and mouth piece. Accessories are available in spe- cialist shop.

description article number EAN

mesh cap 9730002017 4018674000215

PVC adult mask 9730002012 4018674000161

PVC children mask 9730002011 4018674000154

silikon adult mask 9730002014 4018674000185

silikon children mask 9730002013 4018674000178

mouth piece 9730002016 4018674000208

connecting piece/ adapter 9730002015 4018674000192

Page 15 of the Geratherm Micromesh user manual — Troubleshooting
Page 15 · click to enlarge
Page 15

Troubleshooting

What happened Causation Correction

The batteries have run out of power.

Replace the bat- teries.

Clean the mesh cap. If the problem persists, please replace with a new mesh cap.

The indicator turns red.

The mesh cap is dirty.

The device does not work when it is turned on

Check the batter- ies and replace them if neces- sary. Pay atten- tion to polarity.

No power.

Fill medication cup and turn on the nebulizer.

The medication cup is empty.

The indicator is green but no aerosolis produced or the delivery is slower than usual.

The mesh cap is broken.

Replace with a new mesh cap.

The medication cup or the mesh cap is dirty.

Clean those parts or replace mesh cap.

Check the batter- ies and replace them if neces- sary. Pay atten- tion to polarity.

There is medication in the medication cup and the ON/OFF but- ton has been pressed, but no aerosols are being produced.

No power supply.

Check if the mesh cap is cov- ered with liquid. If yes, shake off the liquid and then try again.

The mesh is covered with liquid.

Page 16 of the Geratherm Micromesh user manual — Technical Data
Page 16 · click to enlarge
Page 16

Technical Data

Dimensions 76 mm x 60 mm x 135 mm

Power source 2 x AA alkaline batteries (1.5 V)

Battery life Over 7 hours operating

Power consumption 1.5 W

Vibration frequency 185.5 kHz

Nebulization rate > 0.25 ml/min

Medication capacity 6 ml

Operating conditions 10 °C to 40 °C 30 to 85 % relative humidity

Storage conditions -25 °C to 70 °C up to 93 % relative humidity

IP code (protection rating)IP22

Automatic function Automatic power shut-off if system idles for 15 minutes

Power consumption Normal modeless than 1.5 W Standby mode: less than 0.3 mW

Quality standard The device complies with the Council Directive 93/42/EEC concerning medical devices. A quality management system certifi cated under DIN EN ISO 13485 guarantees compli- ance with the required quality criteria. EN 13544-1 respiratory therapy equipment – part 1: neb- ulizing systems and their components

Distribution of particle size (with mask and mouth piece):

0 5 10 15 20

Show 9 more labels / short text on this page
Weight 190 g · 1,0 · Volumenanteil/ · Volume fraction · 0,8 · 0,5 · 0,3 · 0,0 · Partikelgröße/Particle size [μm]
Page 17 of the Geratherm Micromesh user manual — -25 °C
Page 17 · click to enlarge
Page 17

Symbol Information

Consult instructions for use

Use by/expiry date for masks

Batch code for nebulizer

Manufacturing date of masks

Do not dispose of as household waste

+70 °C

Store between -25 °C and +70 °C

-25 °C

Store between 10 % and 93 % R.H.

Keep dry.

Show 6 more labels / short text on this page
Symbol Referent · Manufacturer · Serial number · Type BF Equipment · 93 % · 10 %
Page 18 of the Geratherm Micromesh user manual — Limited Warranty
Page 18 · click to enlarge
Page 18

Disposal

Adhere to the applicable regulations when disposing of the device and packaging. This product must not be disposed of in domestic waste. All users are obliged to hand in all electrical or electronic devices, regardless of whether or not they contain toxic substances, at a municipal or commercial collection point so that they can be disposed of in an environmentally ac- ceptable manner.

Please remove the batteries before disposing of the device/ unit.

Do not dispose of old batteries with your house- hold waste, but at a battery collection station at a recycling site or in a shop.

Limited Warranty

This ultrasonic nebulizer is guaranteed for 3 years from the

date of purchase. If your nebulizer does not function properly due to defective parts or assembly we will repair it free of charge. All parts are covered by this warranty, except accessories which are subject to normal wear and tear. The warranty does not cover damage to your unit due to improper handling.

Page 19 of the Geratherm Micromesh user manual — Warranty Card
Page 19 · click to enlarge
Page 19

Warranty Card

The Geratherm® micromesh nebulizer is guaranteed for

three years from the date of purchase against manufactur- er’s defect if it is returned to the dealer from whom it was purchased. During this warranty period the unit will be repaired or re- placed free of charge if the defect is due to faulty design or manufacture. This guarantee does not cover any damage or malfunction-

ing that is caused through improper handling, through use otherwise than in accordance with the instructions in this manual or through unauthorised maintenance attempts. Your local dealer cannot declare this manufacturer’s war- ranty null and void, but may extend it by additional guaran- tees at the discretion and at the expense of the dealer.

Model no.:

Serial no.:

Date of purchase:

Dealer’s signature:

Invoice no. / Receipt no.:

Dealer’s name and address:

Geratherm Medical AG Fahrenheitstraße 1 D-98716 Geschwenda Germany Tel: +49 36 205 980

Fax+49 36 205 116 www.geratherm.com
Page 20 of the Geratherm Micromesh user manual — F
Page 20 · click to enlarge
Page 20

Inhalateur basé sur la technologie

du tamis à pores microscopiques

GT-7002

Geratherm Micromesh — Inhalateur basé sur la technologie diagram (page 20)
Fig. 6 · Inhalateur basé sur la technologie

F

NOTICE D’UTILISATION

Sources

References & links

Pages referenced in this manual — manufacturer, support and downloads.

Ask about the Geratherm Micromesh

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the Geratherm Micromesh

Be the first to ask — other owners or our team will answer.

Original document

Geratherm Micromesh manual PDF

164 pages·13.3 MB·Inhaler·PDF
Click any page above to open it here
Page 1 / …
100%
Loading pages…