EN
Beurer logo
Official user manual

Beurer JIH50

76 pages·2.8 MB·Baby care·PDF·ENNLJA
Beurer JIH50
Table of Contents 15 sections ENGLISH · Contents · 3. Safety information and warnings · Warning · … ShowHide
Page 2 of the Beurer JIH50 user manual — Notes on handling batteries
Page 2 · click to enlarge
Page 2
  • Always follow your doctor’s instructions regarding the type of medicine to be used and the dose, frequency, and duration of inhalation.
  • Only use medicines that have been prescribed or recommended by your doctor or pharmacist.
  • If the unit does not function correctly or if you begin to feel unwell or feel pain, stop using the unit immediately.
  • During use, hold the unit away from the eyes, as some nebulised medicines may cause damage to the eyes.
  • Do not operate the unit in the presence of flammable gases.
  • Do not use the device near electromagnetic ransmitters.
  • This device is not intended to be used by children or persons with restricted physical, sensory (e.g. insensitivity to pain) or mental abilities or persons lacking in the required experience or knowledge for safe operation of the device, unless supervised or instructed by a person responsible for their safety.
  • Keep children away from packaging materials (risk of suffocation).
  • Do not use attachments that are not recommended by the manufacturer.
  • Do not dip the unit into water and do not use it in wet rooms. Do not allow any liquid to penetrate the unit.
  • Protect the unit from strong impacts.
  • Do not use the unit if it has fallen or been dropped on the floor, been exposed to extreme humidity or is otherwise damaged. In case of doubt, contact our customer service department or your local dealer.
  • The JIH 50 must only be operated using the appropriate Beurer accessories.The use of other accessories can impair therapeutic efficiency and may damage the device.
  • Remove the batteries from the device if you will not be using it for an extended period. Leaking batteries may damage the device.
  • Do not use any liquids that have a viscosity (thickness) greater than 3, as these may damage the mesh irreparably.
  • Do not use any medicines in powder form (even if dissolved).
  • Do not shake the atomiser, as this may cause liquid to leak, and the device may become damaged.

Power pack

  • Do not use any power pack other than the one supplied.
  • Ensure that the mains socket cannot come into contact with liquids.
  • If adapters or extensions are required, these must comply with the applicable safety regulations. The electrical capacity and the maximum capacity specified on the adapter must not be exceeded.
  • After using the power pack, unplug it from the socket.
  • Connect the unit only to the mains voltage listed on the type plate.

General notes

  • Use the unit only: – On humans, – for its intended purpose (aerosol inhalation) and
  • in the manner described in these instructions for use.
  • Any improper use can be dangerous!
  • In acute emergencies, first aid takes priority.
  • Apart from medicines, use only a saline solution.
  • This unit is not intended for commercial or clinical use, but only for individual, private household use!

Before using the unit for the first time

  • Before you use the unit for the first time, remove all packaging materials.
  • Protect the unit against dust, dirt and moisture.
  • Do not use the unit in very dusty environments.
  • Before use, ensure the device and accessories are at room temperature.
  • Switch the unit off immediately if it is defective or malfunctioning.
  • The manufacturer shall not be held liable for damage or injuries caused by improper or incorrect use.
  • Power cuts, sudden faults or other unfavourable conditions may render the unit unusable. You are therefore recommended to keep a spare device or replacement medicine (as agreed with the doctor).
  • The device must be stored away from sources of heat.
  • Do not use the unit in a room in which a spray has previously been used. If this is the case, ventilate the room before commencing treatment.
  • For reasons of hygiene, each user should use their own set of accessories (mesh atomiser, mask and mouthpiece).
  • Store the device in a location protected against environmental influences. The device must be stored in the specified ambient conditions.

Notes on handling batteries

  • Keep batteries out of the reach of children. Children can put batteries in their mouths and swallow them. This can cause severe harm to their health. In this case, consult a physician immediately!
Page 3 of the Beurer JIH50 user manual — 6. Initial use Before using the unit for the first time
Page 3 · click to enlarge
Page 3
  • Normal batteries must not be recharged, heated or thrown into open flame ( risk of explosion ).
  • Leaking batteries may damage the device. If you do not intend to use the device for prolonged periods, remove the batteries from the battery compartment.
5. Description of device and accessories Overview 1 Cover 2 Medicine container 3 Mesh atomiser 4 Operation LEDs: blue: device ready orange: change batteries 5 On/off button 6 Casing 7 Battery compartment 8 Power pack connec-
Beurer JIH50 — Repairs diagram (page 3)
Fig. 1 · Repairs

Warning

Leaking or damaged batteries can cause burns if they touch the skin. When removing such batteries, wear suitable protective gloves.

Repairs

  • Never open or attempt to repair the unit yourself, as otherwise proper function is no longer guaranteed. Failure to observe this regulation shall void the warranty.
  • If you need to have the unit repaired, contact customer service or an authorised dealer.
4. Useful information on the device Accessories The safe functioning of the device can only be guaranteed if the accessories recommended by the manufacturer are used.

Protecting against leaks When filling the medicine container with medicine, ensure that you do not exceed the maximum mark (8 ml). The recommended filling amount is between 2 and 8 ml. During use, you can tilt the device up to a maximum angle of 45 ° in any direction, without impairing its neb- ulising function or the success of the treatment. How- ever, spraying only takes place if the substance to be sprayed is in contact with the mesh. If it is not, spray- ing stops automatically. However, please try to hold the device as upright as possible.

9 Adult mask 10 Child mask 11 Mouthpiece 12 Connecting piece 13 Power pack Storage bag

Automatic switch-off The device has an automatic switch-off function. To prevent damage to the mesh, the device switches off automatically when the medicine/liquid has been al- most entirely used up. Do not use the device if the medicine container or wa- ter tank is empty. The device switches off automatically if the substance to be sprayed is no longer in contact with the mesh.

6. Initial use Before using the unit for the first time

  • The atomiser and accessories must be cleaned and disinfected before they are used for the first time. For information about this, see “Cleaning and disinfection” on page 16.

tor

Page 4 of the Beurer JIH50 user manual — Repairs
Page 4 · click to enlarge
Page 4
  • Assembly Remove the unit from the packaging.
  • If not already fitted, fit the mesh atomiser (3) onto the neck of the container (2), and lock into place by turning it clockwise until the silver-coloured contact electrodes are facing downwards.
  • Place the container with fitted mesh firmly onto the housing (6) or base of the device, so that you hear it click into place on the left and right.
  • Insert the batteries.
  • This is sufficient for 5 – 9 uses of saline solution (depending on the dose and duration of each use).
  • Operating the device with the power pack Only connect the power pack (13) to the mains voltage listed on the type plate.
  • Open the battery compartment lid to connect the power pack (13).
  • Insert the power pack connector (8) into the connection socket provided (15), and the power pack (13) fully into a suitable socket.
Beurer JIH50 — Repairs diagram (page 4)
Fig. 2 · Repairs
  • Ensure that there is a socket near to where the unit will stand.
  • Lay the mains cable in such a way that no one can trip over it.
  • To disconnect the nebuliser from the mains after use, first switch off the device and then remove the plug from the socket.
  • Do not remove the batteries when operating the device with the power pack (13).
  • The power pack (13) cannot be used for charging batteries.
Beurer JIH50 — Repairs diagram (page 4)
Fig. 3 · Repairs
7. Operation 1. Preparing the atomiser • For reasons of hygiene, it is imperative that the atomiser and accompanying accessories are cleaned and disinfected after every use. If the therapy requires
  • Inserting the batteries
  • To open the battery compartment (7) on the underside of the device, apply light pressure and pull the battery compartment cover in the direction of the arrows, then fold the lid up.
  • Insert four batteries (type AA LR6).
  • Ensure that the poles of the batteries are touching their respective poles in the device (see illustration on the bottom of the casing).
  • Fold the battery compartment lid down again and push it back in place, in the opposite direction to the arrows.
Beurer JIH50 — Repairs diagram (page 4)
Fig. 4 · Repairs

that various different medicines be inhaled in suc- cession, ensure that the atomiser is rinsed after every use. For details, see Cleaning and disinfec- tion on page 16. • Remove the cover (1) from the device.

  • 2 . Filling the atomiser
  • Open the medicine container (2) by opening the clip, and fill the medicine container with an isotonic saline solution or medicine. Do not overfill! The maximum recommended fill amount is 8 ml.
  • Use medication only on the advice of a physician, and check the appropriate inhalation duration and quantity for your needs.
  • If the recommended dose of the medicine is less than 2 ml, top up this quantity to at least 4 ml using only isotonic saline solution. Viscous medicines may also require dilution. Always follow the instructions of your doctor.
  • When changing batteries, ensure that the medicine container is completely empty, otherwise there is a risk of leaks.
  • When the orange LEDs (4) light up, replace all four batteries, because spraying becomes much less effective when the batteries are weak.
  • You can operate the device for approx. 180 minutes with new alkaline batteries.
Page 5 of the Beurer JIH50 user manual — Warning
Page 5 · click to enlarge
Page 5
3. Closing the atomiser • Close the lid of the medicine container (2) and click the clip in place, and then refit the cover (1).
  • 6. End of inhalation
  • After treatment, switch off the device using the on/ off button (5).
  • The LEDs (4) go out.
  • Once the inhalant has been fully sprayed out, the device switches off automatically. For technical reasons, a small amount may still re-
  • 4. Fitting accessories
  • When the required accessories (mouthpiece, adu lt mask or child mask) have been attached to the device, attach the device firmly to the assembled medicine container (2).
  • The adult mask and child mask must be fitted using the connecting piece (12) (adapter).
  • Now place the device to your mouth and close your lips firmly around the mouthpiece. When using either of the masks, place it over the nose and mouth.
  • Start the device using the on/off button (5).
  • The spray mist coming from the device and the illuminated blue LEDs (4) show that the device is working perfectly.
  • main in the medicine container (2).
  • Do not use it.
  • If applicable, disconnect the power pack (13) from the mains.
7. Clean the device For information about this, see “Cleaning and disinfection” on page 16.

8. Cleaning and disinfection

Warning

5. Inhaling correctly • Breathing technique The right breathing technique is important to en-

Observe the following hygiene instructions in order to avoid any health risks. • The nebuliser and accessories are intended for multiple use. Please note that cleaning and hygien- ic reconditioning requirements differ according to the different areas of application.

  • sure that the particles are distributed as widely as possible into the air ways. So that they can enter the air ways and lungs, you must inhale slowly and deeply, hold your breath briefly (5 to 10 seconds) and then exhale quickly.
  • You should only use nebulisers to treat respiratory disorders if advised to do so by your doctor, who will recommend the medicine, dose and application for inhalation therapy.
  • Certain medicines are only available on prescription.
  • Do not clean the mesh or accessories with brushes or similar, as this may cause irreparable damage, and this will mean that the intended result of the treatment is no longer guaranteed.
  • For additional requirements regarding the necessary hygienic preparations (hand washing, handling of medicines or inhalation solutions) in highrisk groups (e.g. cystic fibrosis patients), contact your doctor.

You should hold the device as upright as possi- ble. Tilting it slightly will not, however, affect its function, since the container is leakproof. To en- sure that the device functions correctly, make sure that you do not tilt it more than 45 ° in any direction and that the medicine is in contact with the mesh while in use.

  • Cleaning The mesh atomiser and container, as well as accessories such as mouthpiece, mask, connecting piece, etc. must be cleaned with hot water after each use. Carefully dry all components using a soft cloth. Once all components are fully dry, reassemble the atomiser and place the components in a dry, sealed container or carry out disinfection.
  • Remove the mouthpiece and any accessories from the atomiser.
  • Add 6 ml of clean water to the medicine container (2) and switch on the device to first remove any medicine residue from the mesh. (Please spray the entire liquid).
  • Remove the batteries.
  • Remove the detachable parts from the device, e.g. mesh atomiser or medicine container and cover.

Essential oils, cough medicines, solutions designed for gargling, and drops for application to the skin or for use in steam baths are wholly unsuitable for inhalation using a nebuliser. These substances are often highly viscous and can impair the correct function of the de- vice and hence affect the effectiveness of its applica- tion in the long term. In the case of oversensitivity of the bronchial system, medicines containing essential oils can sometimes cause acute bronchiospasm (a sudden, cramp-like tightening of the bronchi accompanied by breathless- ness). Always ask your doctor or pharmacist for advice!

Page 6 of the Beurer JIH50 user manual — 8. Cleaning and disinfection
Page 6 · click to enlarge
Page 6

Disassembly To continue with cleaning its individual components, disassemble the atomiser (2, 3): • Remove the medicine container (2) and mesh atomiser (3) by pressing both brackets inwards and drawing the unit out.

onto the unit. The unit must be operated only when it is completely dry. • Do not spray any liquids into the ventilation slots. Any liquid that penetrates into the unit can dam- age the electrical parts or other components of the nebuliser and impair the function of the device.

Beurer JIH50 — 8. Cleaning and disinfection diagram (page 6)
Fig. 5 · 8. Cleaning and disinfection

Disinfection Follow the steps listed below carefully in order to disin- fect your atomiser and the accessories. It is advisable to disinfect the individual components after the last time they are used each day at the latest. You can disinfect the JIH 50 and its accessories in two different ways, either with ethyl alcohol or in boil- ing water. First clean the atomiser and the accessories as de- scribed under “Cleaning”. You can then continue with disinfection.

  • Remove the mesh atomiser (3) from the medicine container (2) by turning it anticlockwise and drawing it out.
  • The nebuliser is later reassembled in the reverse order.

Ensure that you do not touch the mesh (3), as this may destroy it.

  • Disinfection with ethyl alcohol
  • Pour 8 ml ethyl alcohol into the medicine container (2). Close the container. Leave the alcohol in the container for at least 10 minutes.
  • For better disinfection, shake the unit gently from time to time.
  • Afterwards, pour the ethyl alcohol out of the container.
  • Repeat this process, using water this time.
  • Place the medicine container (2) in a position that allows you to apply a few drops of ethyl alcohol to the mesh (3). Allow the alcohol to work for 10 minutes.
  • Finally, clean all parts again under running water.
  • Never hold the complete device under running water to clean it. When cleaning, ensure that all residues are removed. Never use any substances that may be poisonous when in contact with the skin or mucous membranes, or when swallowed or inhaled. If required, you can clean the device casing with a lightly moistened cloth that has been soaked in mild soapy water.
  • If there is any residue from medical solutions or contamination on the silver-coloured contacts of the device or atomiser, clean with a cotton bud moistened with ethyl alcohol.
  • Clean the outside of the mesh and outside of the medicine container with a cotton bud dampened with ethyl alcohol.
  • Disinfection with boiling water
  • The atomiser should be disassembled as described and, together with the mouthpiece, placed in boiling hot water for 15 minutes. Avoid the parts coming into contact with the hot base of the pot.
  • You can also disinfect the atomiser and the mouthpiece with a standard vaporiser. Observe the manufacturer’s instructions for use for the vaporizer when doing this.
  • The atomiser must not be placed in the microwave.
  • Before you clean the unit, always switch it off, unplug it and let it cool off.
  • Do not use abrasive detergents and never immerse the unit in water.
  • Ensure that no water penetrates inside the unit!
  • Do not put the device or accessories in a dishwasher.
  • Do not touch the unit with wet hands while it is plugged in; do not allow any water to be sprayed

The face masks and air hose must not be placed in hot water! • We recommend disinfecting the masks using a re- tail disinfectant.

Page 7 of the Beurer JIH50 user manual — Troubleshooting
Page 7 · click to enlarge
Page 7
  • Drying
  • Carefully dry all components using a soft cloth.
  • Shake the mesh atomiser (3) lightly backwards and forwards (5 to 10 times), so that the water inside the mesh is removed from the small holes.
  • Place the individual components on a dry, clean and absorbent surface and allow to dry completely (for at least 4 hours).

Output is too low.

1. The batteries are flat. Replace the batteries or connect the power supply unit and try again. 2. There are air bubbles inside the medicine container which are pre- venting the medicine from coming into continuous contact with the mesh. Please check and remove any air bubbles. 3. Particles on the mesh are imped- ing output. To remove the par- ticles, fill the nebuliser with 2 to 3 drops of vinegar and 3 to 6 ml of water, then nebulise this solu- tion completely. Do not inhale this spray and ensure that you clean and disinfect the medicine con- tainer afterwards (see page 16). If the output does not improve, re- place the mesh. 4. The mesh is worn out. Which medi- cines are suit- able for inhala- tion?

Ensure that all parts are thoroughly dried after cleaning, otherwise there is an increased risk of bacterial growth. Once all components are fully dry, reassemble the nebuliser and place the components in a dry, sealed container.

Ensure that the mesh atomiser (3) has dried completely by shaking it. Otherwise, nebulisation may not function properly after reassembling the device. If this is the case, disassemble the mesh atomiser (3) again and shake it so that the water can escape. Once it has been reassembled, the nebuliser should function properly again as usual.

Only the doctor can advise you which medicine to use to treat your condition. Consult your doctor. With the JIH 50, you can atomise medicines with a viscosity lower than 3. Some inhala- tion solution remains in the nebuliser.

  • Material resistance
  • When choosing a cleaning or disinfecting agent, note the following: Only use mild cleaning or disinfectant agents in the quantities recommended by the manufacturer.
  • As with all plastic components, frequent use and cleaning of the atomiser and the accessories can lead to a certain amount of wear. Over time, this can alter the aerosol properties and may eventually affect the efficiency of therapy. We therefore recommend that you replace the atomiser and accessories at least once a year.

This is normal and occurs for techni- cal reasons. Stop inhaling as soon as you can hear a marked differ- ence in the sound made by the at- omiser, or when the device switches off automatically due to insufficient inhalant. What special steps should be taken for babies and children?

1. In babies, the mask should cover the mouth and nose to guarantee effective inhalation. 2. In children, the mask should also cover the mouth and nose. Nebulisation next to a sleeping person is not suitable because insufficient medicine reaches the lungs. Note: Children should only use the device with help and under supervision of an adult. Never leave a child alone with the nebuliser. Inhalation with the mask takes longer.
  • Storage
  • Do not store the nebuliser in a damp atmosphere (e.g. in the bathroom) or transport it together with damp objects.
  • Store and transport the nebuliser away from direct sunlight.

Troubleshooting

The nebuliser produces little or no aerosol.

1. Insufficient medicine in the atomiser. 2 . The atomiser is not held in an upright position. 3 . The medicine used is unsuitable for spraying (e.g. too thick, viscosity must be less than 3). The medicine solution should be

This is for technical reasons. Less medicine is inhaled per breath through the mask holes than us- ing the mouthpiece. The aerosol is mixed with ambient air through the holes in the mask. Does each user need their own accessories?

This is absolutely necessary for hy- giene reasons.

specified by the doctor.

Show 4 more labels / short text on this page
Problem/ Question · Possible Cause/Remedy · Problem/ Question · Possible Cause/Remedy
Page 8 of the Beurer JIH50 user manual — PP, PVC, EVA, PET
Page 8 · click to enlarge
Page 8
9. Technical specifications Model JIH50 Type IH50 Dimensions (LxWxH) 72 x 59 x 143 mm

This diagram may not be applicable for suspensions or highly viscous medicines. More information can be ob- tained from the relevant medicine manufacturer.

If the unit is used outside of the specifications, proper function is no longer guaranteed! We reserve the right to make technical changes to improve and further de- velop the product. This device and its accessories comply with European standards EN60601-1 and EN60601-1-2, as well as EN13544-1, and is subject to special safety measures in terms of electromagnetic tolerance. Note that porta- ble and mobile RF communication equipment can af- fect this unit. More details can be requested from the stated Customer Service address or found at the end of the instructions for use. The unit conforms to the re- quirements of the European Directive for Medical Pro- ducts 93/42/EC, the MPG (German Medical units Act).

Weight 238 g incl. batteries Battery operation 4 x 1.5 V type AA, Mignon (LR6) Volume max. 8 ml Medicine flow approx. 0.25 ml/min. Oscillation frequency 100 kHz Housing material ABS / TPE Mains connection 100 – 240 V ~, 50 – 60 Hz; 0,15 A Operating conditions

Temperature+5 °C – +40 °C Relative humidity: <80 % non-condensing Storage and transport conditions
Temperature-10 °C – +45 °C Relative humidity: <80 % non-condensing
  • ELECTROMAGNETIC COMPATIBILITY
  • The device complies with current specifications with regard to electromagnetic compatibility and is suitable for use in all premises, including those designated for private residential purposes. The radio frequency emissions of the device are extremely low and in all probability do not cause any interference with other devices in the proximity.
  • It is recommended that you do not place the device on top of or close to other devices. Should you notice any interference with other electrical devices, move the unit or connect it to a different socket.
  • Radio equipment may affect the operation of this device.

Subject to technical modifications. Wearing parts are not covered by the guarantee.

Spare parts

Designation Material REF Mouthpiece PP 162.658 Adult mask PVC 162.712 Child mask PVC 162.659 Mesh atomiser with medicine container

PC 162.711

Power pack PC+ABS 071.18 Connecting piece PC 162.665 Yearpack (Mouthpiece, Adult mask, Child mask, Mesh atomiser, Connecting piece)

PP, PVC, EVA, PET

10. Disposal The appliance should be disposed of accord- ing to Regulation 2002/96/EC – WEEE (Waste Electrical and Electronic Equipment). In case of queries, please contact the municipal authorities responsible for waste disposal in your area. Used batteries do not belong in the household waste. You are legally obliged to dispose of the batteries. Dis- pose of them via your specialist electrical supplier or local collecting point for recyclable waste. Note: Batteries containing pollutant substances are marked as follows: Pb = Battery contains lead, Cd = Battery contains cadmium, Hg = Battery contains mercury. The batteries for this device do not contain any toxic substances.

Beurer JIH50 — PP, PVC, EVA, PET diagram (page 8)
Fig. 6

All measurements were obtained with a sodium chlo- ride solution using a laser diffraction method.

Fix-it

Troubleshooting

Common problems, likely causes, and what to do.

ProblemPossible causeRemedy
Output is too low.1. The batteries are flat. Replace the batteries or connect the power supply unit and try again. p.7
Output is too low.2. There are air bubbles inside the medicine container which are pre- venting the medicine from coming into continuous contact with the mesh. Please check and remove any air bubbles. p.7
Output is too low.3. Particles on the mesh are imped- ing output. To remove the par- ticles, fill the nebuliser with 2 to 3 drops of vinegar and 3 to 6 ml of water, then nebulise this solu- tion completely. Do not inhale this spray and ensure that you clean and disinfect the medicine con- tainer afterwards (see page 16). If the output does not improve, re- place the mesh. p.7
Output is too low.4. The mesh is worn out. p.7
Which medi- cines are suit- able for inhala- tion?Only the doctor can advise you which medicine to use to treat your condition. Consult your doctor. With the JIH 50, you can atomise medicines with a viscosity lower than 3. p.7
Some inhala- tion solution remains in the nebuliser.This is normal and occurs for techni- cal reasons. Stop inhaling as soon as you can hear a marked differ- ence in the sound made by the at- omiser, or when the device switches off automatically due to insufficient inhalant. p.7
What special steps should be taken for babies and children?1. In babies, the mask should cover the mouth and nose to guarantee effective inhalation. 2. In children, the mask should also cover the mouth and nose. Nebu- lisation next to a sleeping person is not suitable because insufficient medicine reaches the lungs. Note: Children should only use the device with help and under supervi- sion of an adult. Never leave a child alone with the nebuliser. p.7
Inhalation with the mask takes longer.This is for technical reasons. Less medicine is inhaled per breath through the mask holes than us- ing the mouthpiece. The aerosol is mixed with ambient air through the holes in the mask. p.7
Does each user need their own accessories?This is absolutely necessary for hy- giene reasons. p.7
The nebuliser produces little or no aerosol.1. Insufficient medicine in the atomiser. p.7
The nebuliser produces little or no aerosol.2. The atomiser is not held in an up- right position. p.7
The nebuliser produces little or no aerosol.3. The medicine used is unsuitable for spraying (e.g. too thick, viscos- ity must be less than 3). The medicine solution should be specified by the doctor. p.7

Ask about the Beurer JIH50

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the Beurer JIH50

Be the first to ask — other owners or our team will answer.

Original document

Beurer JIH50 manual PDF

76 pages·2.8 MB·Baby care·PDF
Click any page above to open it here
Page 1 / …
100%
Loading pages…