Instruction/ maintenance manual of the product MP22A UltraViolet Devices
Go to page of 20
375 Marcus Boulevard • Hauppauge, NY 1 1 788 • USA 63 1 .273.0500 • Fax: 63 1 .273.077 1 e-mail: info@ultraviolet.com Extensive Product Information A vailable at: Document No. 98 - 1 123E1 • Revised February 2007 • © 1994-2007 Atlantic Ultraviolet Corporation www .
2 T ABLE OF CONTENTS S AFETY W ARNINGS .....................................................................................................................................................3 SAFETY INSTRUCTIONS .........................................
3 SAFETY W ARNINGS • All personnel should be alerted to the potential hazards indicated by the product safety labeling on this unit. • The following conventions are used to indicate and classify precautions in this manual and on product safety labeling.
4 PRODUCT APPLICA TION C ONSTRUCTION • The water purifier is designed to mount horizontally . • A drain port on the chamber aids in draining of the purifier . • The water purifier chamber is passivated and electropolished type 316 Stainless S teel.
5 INST ALLA TION L OCA TION 1. The water purifier is intended for indoor use only . The water purifier should be protected from the elements and from temperatures below freezing. The ambient temperature, in the area surrounding the water purifier , should be between 33 F and 100 F.
6 6 3. The water purifier should be mounted horizontally on a flat dry surface. Secure the water purifier using the mounting holes in the ballast housing or with the optional wall mounting kit. The purifier should not be solely supported by its plumbing connections.
7 R ECOMMENDED O PTIONS 1 . Guardian ™ Ultraviolet Monitor: V isually indicates the level of germicidal ultraviolet energy that penetrates the quartz sleeve and the water within the water purifier . The ultraviolet monitor is capable of operating an optional audio alarm and/or solenoid valve.
8 MAINTENANCE The water purifier is designed to operate with a minimal amount of maintenance, providing the water quality does not exceed maximum concentration levels, see “ W ater Quality ” in the “ Pr oduct Application ” section.
9 S tep 2 - Remove End-Cap S tep 3 - W ithdraw Lamp S tep 4 & 5 - Remove Sockets S tep 6 - Remove Lamp L AMP I NST ALLA TION OR R EPLACEMENT Figure 3 - Lamp Replacement 1. Disconnect power to water purifier . 2. Remove both easy-off end caps by pulling each cap of f static gland nut.
10 1. Disconnect power to water purifier . 2. Shut off water supply to water purifier via inlet and outlet shut of f valves. Drain chamber by removing drain plug.
11 R EPLACEMENT OF B ROKEN Q UARTZ S LEEVE CAUTION: Broken Quartz is SHARP . It is r ecommended that protective goggles and gloves are worn when handling br oken quartz sleeves. 1. Disconnect power to water purifier . 2. Shut off water supply to the water purifier via the inlet and outlet shut of f valves.
12 O PTIONAL G UARDIAN ™ ™ U L TRA VIOLET M ONIT OR S ENSOR P ROBE C LEANING Figure 5 - Ultraviolet Monitor and Sensor Probe If after cleaning the quartz sleeve, there is no significant improvement in the ultraviolet intensity , as shown on the intensity meter , it may be necessary to clean the Ultraviolet Monitor's sensor probe.
13 Problem Possible Cause Corrective Action No electrical power… V erify that the purifier is connected to a live power source. Ballast Disconnect the purifier from utility power . Replace lamp. Lamp-out Protection The purifier must remain disconnected from power for (MP16A, MP22A & MP36C) two (2) minutes to reset.
14 TECHNICAL SPECIFICA TIONS T able 4 - T echnical Specifications Model: MP13A MP16A MP22A MP36C MP49C Flow Rate (GPM): 2 3 6 12 20 InletOutlet Size: 1 / 2 " m NPT 3 / 4 " m NPT 3 / 4 " m NPT 1 " m NPT 1- 1 / 2 " m NPT Number of Lamps: 1 1 1 1 1 Lamp Model No.
15 OPTIONAL ACCESSORIES Optional accessories are available for operation at 120v 60Hz or 220v 50Hz, unless specified. Please specify . Consult Factory for 12v DC or other special input power requirements.
16 REPLACEMENT P AR TS Mighty Pure ™ ™ Models MP13A, MP16A & MP22A Figure 6 - Exploded V iew MP13A, MP16A & MP22A.
17 ITEM MP13A MP16A MP22A NO. DESCRIPTION QTY P AR T NO. QTY P AR T NO. QTY P AR T NO. 1 Easy-Off End Cap 2 25-1499A1 2 25-1499A1 2 25-1499A1 2 Rubber O-ring, Static Gland 2 00-1 108B 2 00-1108B 2 00-.
18 REPLACEMENT P AR TS Mighty Pure ™ ™ Models MP36C & MP49C Figure 7 - Exploded V iew MP36C & MP49C.
19 ITEM MP36C MP49C NO. DESCRIPTION QTY P AR T NO. QTY P AR T NO. 1 Easy-Off End Cap 2 25-1499A1 2 25-1499A1 2 Rubber O-ring, Static Gland 2 00-1108B 2 00-1 108B 3 Lamp Spacer - - 2 25-0209A 4 Lead W .
20 W ARRANTY W e warrant that this product will be free from defects in material and workmanship for a period of one year from the date of shipment thereof or the product’ s total rated life, whichever first occurs.
An important point after buying a device UltraViolet Devices MP22A (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought UltraViolet Devices MP22A yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data UltraViolet Devices MP22A - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, UltraViolet Devices MP22A you will learn all the available features of the product, as well as information on its operation. The information that you get UltraViolet Devices MP22A will certainly help you make a decision on the purchase.
If you already are a holder of UltraViolet Devices MP22A, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime UltraViolet Devices MP22A.
However, one of the most important roles played by the user manual is to help in solving problems with UltraViolet Devices MP22A. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device UltraViolet Devices MP22A along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center