Instruction/ maintenance manual of the product TIC-LF414C Toshiba
Go to page of 16
TIC-LF414C Introduction The electromagnetic flowmeter uses Faraday’s Law of electromagnetic induction to measure the process flow. The device consists of two units: a detector, through which the flu.
TIC-LF414C 2 Power consumption: 17W(27VA) or less 19W(29VA) or less (with PROFIBUS) Conformance to European Co mmunity Directives: EMC directive 89/336/EEC The low voltage 93/68/EEC PED 97/23/EC (Note 1) Note : See Table 1 for detail.
TIC-LF414C 3 Digital input DI (opt.) Signal type: 20 to 30Vdc voltage signal Input resistance: 2.7k Ω Number of inputs: one point DI function — One of the following functions can be assigned to the optional DI signal. Range switching — Selects either the higher or lower range in the unidirectional or bidirectional 2-range setting.
TIC-LF414C 4 Conditions when power fails: Parameter setting values are stored in non–volatile memory and the values will be restored when the power r eturns to normal condition.
TIC-LF414C 5 Installation Dimensions Unit : inch( mm) L1 Φ D1 (L2 ) 2. 5 2 ( 64 ) 5.63( 143) 1.42 ( 36) 8.86 ( 225) 5.16( 131) ANSI class 150 and class 300 dimensions: Meter size (inch) L1 (inch) L2 (inch) D1 (inch) Weight (lb) 1/2 2.76 9.92 1.93 approx.
TIC-LF414C 6 Unit : inch(mm) (L2 ) L1 8.86( 225) 5.63 ( 143) 1.42( 36) 5.16( 131 ) Φ D1 Note1: Eye bolts are provided at the top fo r flow mete rs sized 8"(200mm). Note2: 1 inch = 25.4mm ANSI class 150 and class 300 dimensions Meter size (inch) L1 (inch) L2 (inch) D1 (inch) Weight (lbs) 1-1/2 3.
TIC-LF414C 7 Unit : inch(mm) R ( P T ) 1/2Male S c r e w ( T w o Pl a c e s) 3.46( 88) 1.57 ( 40 ) 4.96 ( 126 ) 3.58 ( 91 ) 1.38( 35) (L2 ) Φ 1.93( Φ 49) 2.52( 64 ) L1 Note1: Cable glands are not provide d for LF414 of FM and CSA appr oved type. Refer to the part of Cable connection port at detector.
TIC-LF414C 8 Unit : inch(mm) 3.46(88) 1.57 ( 40 ) 4.96(126) 1.38(35) 3.58(91) R(PT)½ Mal e Screw (T wo places) (L2) Φ 2.60( Φ 66) 3.43 ( 87 ) L1 Note1: Cable glands are not provide d for LF414 of FM and CSA appr oved type. Refer to the part Cable c onnection port at detector.
TIC-LF414C 9 Unit : inch(mm) 3.46( 88) 1 .57 ( 4 0 ) 4.9 6( 126) 1. 38 ( 3 5 ) 3.58(91) R( PT )½M a le Sc r e w ( T w o p l a c e s ) (L 2) L1 Φ D1 Note1: Cable glands are not provide d for LF414 of FM and CSA appr oved type. Refer to the part of Cable connection port at detector.
TIC-LF414C 10 矢視 A 信号ケ ー ブ ル グラ ン ド 電源ケ ー ブ ル グ ラ ン ド 封止栓 励磁 ケ ー ブ ル グ ラ ン ド 電流出力ケ ー ブ ル グ ラ ン ド 接地端子 (M 4 ) 取付金具 4 - φ 11 取付金具 82 (3 6 ) 1 5 6 1 3 1 51 178 20 0 1 51 74 ± 0 .
TIC-LF414C 11 ・ Separate type LF410/LF612 and LF414/LF612F flowmeters T erminal board Connected detecto r IV wire 5.5 mm 2 or more Power switch (External double- pole power switch ) Thick walled ste.
TIC-LF414C 12 Wiring Precautions (1) Explosion proof type flow meters are not provided cable glands. Refer to the part Cable connection port at detector and converter. (2) Connect the grounding wire (IV wire 5.5mm² or more) to a good earth ground (100 Ω or less ground resistance).
TIC-LF414C 13 Meter Size To select the meter size: See Table 3 to 4 and find meter sizes within the velocity of 0.3 to 32.8 ft/s (0.1 to 10m/s) for a specified full-scale (measuring range high limit) flow. Select one that has its full-scale velocity between 3.
TIC-LF414C 14 Table 6. Specification Code (Waf er type detector LF410 Series) Detector ca tegory Model Specific ation Code Normal type Ex. type 1 2 3 4 5 6 7 8 9 10 11 12 13 14 Description Gr.
TIC-LF414C 15 Table 7. Specification Code for converter s Model Specific ation Code 1 2 3 4 5 6 7 8 9 10 11 12 13 14 Contents L F 6 1 Elec tromagnetic flowmeter c onverter LF610 type LF612 type 0 Comb.
TIC-LF414C TIC-LF414C Specifications are subject to change without notice. Printed in Japan 2008-6 (TDOC) © TOSHIBA Corporation 2008 All Rights Reserved. http://www.toshiba.com/ind / Misuse of this product can result in damages to property or hum an injury.
An important point after buying a device Toshiba TIC-LF414C (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Toshiba TIC-LF414C yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Toshiba TIC-LF414C - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Toshiba TIC-LF414C you will learn all the available features of the product, as well as information on its operation. The information that you get Toshiba TIC-LF414C will certainly help you make a decision on the purchase.
If you already are a holder of Toshiba TIC-LF414C, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Toshiba TIC-LF414C.
However, one of the most important roles played by the user manual is to help in solving problems with Toshiba TIC-LF414C. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Toshiba TIC-LF414C along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center