Instruction/ maintenance manual of the product MX300-I Teledyne
Go to page of 48
Portable Oxygen Monitor Teledyne Analytical Instruments i OPERA TING & SER VICE INSTRUCTIONS FOR MX300-I P ORTABLE O XYGEN M ONITOR 0086 TYPE B EQUIPMENT: Equipment providing a particular degree o.
MX300-I Teledyne Analytical Instruments ii Copyright © 2002 Teledyne Analytical Instruments All Rights Reserved No part of t his ma nual may be re produced, t ransmitt ed, transcribe d, stored in a r.
Portable Oxygen Monitor Teledyne Analytical Instruments iii About This Manual The MX300-I Operator's Manual pr ovides both introductory and detailed information for configuri ng and operating these instruments. The manual takes you from the tim e you unpack the instrument until you complete the first gas analysis.
MX300-I Teledyne Analytical Instruments iv Safety Messages Your safety and the safety of ot hers are very important. Please carefully read the following safety m essages. Safety message are indented to aler t the user of potential hazards. Each safety message is associated with a safety alert symbol.
Portable Oxygen Monitor Teledyne Analytical Instruments v T able of Content s Safety Me ssages .......................................................................... iv List of Figures...............................................................
MX300-I Teledyne Analytical Instruments vi 2.3.1 Humidity 26 2.3.2 Temperature 27 2.3.3 Pressure 27 2.3.4 Discrepancy in Readings 28 2.3.5 Anesthetic Gases 28 2.3.5.1 Gases That Induce Reading Error 28 2.3.5.2 Care After Use in Nitrous Oxide 29 2.3.6 Cleaning 30 2.
Portable Oxygen Monitor Teledyne Analytical Instruments vii List of Figures Figure 1-1: MX300- I Front View ...................................................... 9 Figure 2-1: Installing the R17MED Sensor .................................... 16 Figure 2-2: Sensor Cable Connection to Monitor .
MX300-I Teledyne Analytical Instruments viii List of T ables Table 2-1: Oxygen Reading Error in a Mixture of Anesthetic Gas . 29 Table 3-1 Troubl eshooting ............................................................ 37 Table 3-2 E rror C odes .......
Portable Oxygen Monitor Introduction Teledyne Analytical Instruments 9 Introduction Teledyne Analytical Instrument s MX300-I Oxygen Monitor with alarms here after referred to as MX 300-I is a portable instru ment that provides fast and accurate oxygen monitoring and incorporates an audio/visual alarm capability.
Introduction MX300-I Teledyne Analytical Instruments 10 Options for alternate display config uration). An integral battery life indicator is displayed “on demand” with the MX300-I. The instrument is powered by th ree AA alkaline batteries and is designed to operate for 2000 hours in non-alarm state.
Portable Oxygen Monitor Introduction Teledyne Analytical Instruments 11 1.2 Features The MX300-I is a compact, versatile instrument capable of rapidly measuring the oxygen content of an atmosphere or environm ent accurately to ±2% over the ra nge 0-100% oxygen.
Introduction MX300-I Teledyne Analytical Instruments 12 Section 2.1.5. In addition to the above instru ment configuration options, the following optional equipment is ava ilable for your instrum ent: .
Portable Oxygen Monitor Introduction Teledyne Analytical Instruments 13 is illuminated on the LCD and the audible and visual alarms are activated whenever a fault is d etected. 1.5.1 Sensor The MX300-I uses the Teledyne Class R17MED disposable oxygen sensor.
Introduction MX300-I Teledyne Analytical Instruments 14 The oxygen level is calculated a nd then displayed on the liquid crystal display (LCD) on the front panel. In the MX300-I monito r, user-programmed hi gh and low alarm set points are stored in random access me mory (RA M), which resi des within the microprocesso r.
Portable Oxygen Monitor Operation Teledyne Analytical Instruments 15 Operation Note: Upon receipt, INSPECT THE ENTIRE UNIT FOR DAMAGE. Check the unit and a ll included accessories for broken or loose parts. If damaged, DO NOT USE. Notify the shipper, and consult Teledyne Analytic al Instruments.
Operation MX300-I Teledyne Analytical Instruments 16 2.1.1 Sensor Installation or Replacement Note: The R17MED oxygen sensor must be installed before the oxygen analyzer/monitor can be operated 1. Remove the new sensor from its protective bag. Inspect the sensor for damage or electrolyte leakage.
Portable Oxygen Monitor Operation Teledyne Analytical Instruments 17 Figure 2-2: Sensor Cable Connection to Monitor Note: When the MX300-I instrument is used for diffusion sampling (i.e., incubators, tent s, etc.), the plastic flow diverter must be removed from the R17MED sensor.
Operation MX300-I Teledyne Analytical Instruments 18 2.1.2 Mounting The MX300-I can be mounted in several ways depending on the optional equipment ordered at the time of purchase.
Portable Oxygen Monitor Operation Teledyne Analytical Instruments 19 2.1.3 Battery Installation Note: Three “AA” alkaline batteries must be installed in the unit before the monito r will operate.
Operation MX300-I Teledyne Analytical Instruments 20 4. Re-install the battery compar tment door. W hen the unit is first turned on the display will m omentarily display all LCD segments. During this period diagnostic tests are being conducted to insure the circu its are functioning correctly.
Portable Oxygen Monitor Operation Teledyne Analytical Instruments 21 made to the settings. If the batt eries have just been installed the LCD will flash 00.0 4. As with most oxygen analyzer (s) the highest level of accuracy is achieved when cal ibration is conducted using 100% oxygen.
Operation MX300-I Teledyne Analytical Instruments 22 Figure 2-8: Setting the Alarms should never be used where a high degree of accuracy is needed. Note: Never calibrate the unit in humidified gas, as water vapor makes the oxygen concentration appear lower than it really is.
Portable Oxygen Monitor Operation Teledyne Analytical Instruments 23 designed to prevent cr ossing of HI/LO alarm settings. If you attempt to set the LO alarm hi gher than the HI alarm , it will push the HI alarm setting up as you continue to rise the LO alarm set point.
Operation MX300-I Teledyne Analytical Instruments 24 Figure 2-9: 0-1 VDC or RS 232 Digital Output Port If your instrument is set fo r analog (0-1 VDC) output, you can reconfigure it to use the digital output by changing a jumper on the internal PC board.
Portable Oxygen Monitor Operation Teledyne Analytical Instruments 25 2.2 Use 2.2.1 Procedure Note: Prior to use, always test the batteries and alarms (if applicable).
Operation MX300-I Teledyne Analytical Instruments 26 R17MED sensor so that it does not in terfere with the rapid exchange of gases to and from the sensing surface of the sensor. CAUTION: FAILURE TO REMOVE THE DIVERTER IN T HESE APPLICATION AREAS WILL RESU LT IN A MARKED LOWERING OF THE RESPONSE TIME OF THE SENSOR.
Portable Oxygen Monitor Operation Teledyne Analytical Instruments 27 2.3.2 T emperature The R17MED oxygen sensor adjusts for ambient temperature changes in the range of 0–40°C (32–106° F).
Operation MX300-I Teledyne Analytical Instruments 28 2.3.4 Discrepancy in Readings The MX300-I instrument should be used to measure the oxygen concentration exiting another oxygen mixing device or life support system (i.e., a blender, incubato r or anesthes ia machine).
Portable Oxygen Monitor Operation Teledyne Analytical Instruments 29 Table 2-1: Oxygen Reading Error in a Mixture of Anesthetic Gas Gas or Vapor Level (Balance: Mixture of 30% O 2 / 70% N 2 O, except .
Operation MX300-I Teledyne Analytical Instruments 30 After exposure to nitrous oxide mixt ures, the sensor should be left in 100% oxygen overnight (e.g., left in a breathing circu it that has been flushed with pure oxygen). If this is not practical, when using the Tee, remove the plastic flow diverter and l eave the sensor in room air.
Portable Oxygen Monitor Operation Teledyne Analytical Instruments 31 • Test batteries regularly and re place when battery indicator shows low battery (no bars remaining) • Make sure keys are locked by using the LOCK/UNLOCK key feature. • Keep the unit, sensor and connections dry, or on the dry side of the breathing circuit.
Operation MX300-I Teledyne Analytical Instruments 32 • Use cleaning agents or liquids in the cable receptacles o r around the battery compartment. • Place the unit itself in a water vapo r-saturated environment. • Expose the LCD to excessive sunlight.
Portable Oxygen Monitor Service Manual Teledyne Analytical Instruments 33 Service Manual 3.1 General Service Information The Teledyne Model MX300-I Portable Oxygen Monitor is designed to be robust yet compact in size.
Service Manual MX300-I Teledyne Analytical Instruments 34 DO: Recalibrate after replacing batteries. The MX300-I instrument incorporates a battery test feature that provides the approximate amount of useful life rem aining on the set of batteries. Excessive alarm activation will wear down the battery faster than usual.
Portable Oxygen Monitor Service Manual Teledyne Analytical Instruments 35 functional at this point. Af ter calibration is completed, the ala rm set points will need to be reset. To calibrate the instrument refer back to Section 2.1.3. 3.6 Alarms To set the alarms on the MX300-I monitor, refer to Section 2.
Service Manual MX300-I Teledyne Analytical Instruments 36 during this period since this error will be eliminated when both the thermistor and sensing electrode have had sufficient time to come to thermal equilibrium.
Portable Oxygen Monitor Service Manual Teledyne Analytical Instruments 37 investigation should be made until the discrepancy in readin gs is resolved. The troubleshooting sec tion of this manual may provide additional assistance in locating the problem.
Service Manual MX300-I Teledyne Analytical Instruments 38 Symptom Why What To Do the reading again. E) Try calibrating with a known good sensor; if this fails, see symptom "Reading drifts over 2–3%..." The sensor does not react to changes in oxygen concentration, or the readings are unstable and drifting.
Portable Oxygen Monitor Service Manual Teledyne Analytical Instruments 39 Symptom Why What To Do ventilate the patient, water vapor actually dilutes the gas. See Appendix: Humidity, Temperature. D) If a blender is used, check its calibration. See Appendix: Discrepancy in Readings.
Service Manual MX300-I Teledyne Analytical Instruments 40 Symptom Why What To Do Cannot adjust calibration or alarm settings Critical settings require two keys be pressed in a specific order. LOCK/UNLOCK key is active A) If display is not flashing press lock key once to activate keys.
Portable Oxygen Monitor Service Manual Teledyne Analytical Instruments 41 50 5 The ADC circuit failed during POST. 6 (5) 6 (There is no indication of which key is stuck) A stuck key has been detected.
Service Manual MX300-I Teledyne Analytical Instruments 42 non-abrasive cleaner solution to periodically clean the keypad and screen. An aerosol jet spray of the type commonly used to clean computer keyboards can be used to dislodge dirt and accum ulations from the keypad.
Portable Oxygen Monitor Appendix Teledyne Analytical Instruments 43 Appendix A.1 S pecifications Range: 0-100% oxygen Accuracy: + 2% of full scale (at constant temperature and pressure) Response Time: 90% in less than 8 seconds at 25 °C Battery Life: Approximately 2000 hr.
Appendix MX300-I Teledyne Analytical Instruments 44 A.2 S p are Part s List QTY PART NO DESCRIPTION 1 C43690-R17MED Micro-Fuel Cell R17MED with flow diverter P/N A50057 3 B99 “AA” size alkaline ba.
Portable Oxygen Monitor Appendix Teledyne Analytical Instruments 45 Note: Orders for replacement parts should include the part number (if available) and t he model and serial num ber of the instrument for which the parts are intended.
Appendix MX300-I Teledyne Analytical Instruments 46 BLANK.
Portable Oxygen Monitor Index Teledyne Analytical Instruments 47 Index accessories, 44 accuracy, 43 address, 42, 45 alarm setpoints, 2 2 alarm setpoints, 22 alarm setpoint s, 10, 14 alarm settings cro.
Index MX300-I Teledyne Analytical Instruments 48 sensor mount ing, 17 set up, 15 setpoints. See alarm setpoints signal output , 23 signal processing , 13 signal processor, 1 2 spare parts listing, 44 .
An important point after buying a device Teledyne MX300-I (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Teledyne MX300-I yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Teledyne MX300-I - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Teledyne MX300-I you will learn all the available features of the product, as well as information on its operation. The information that you get Teledyne MX300-I will certainly help you make a decision on the purchase.
If you already are a holder of Teledyne MX300-I, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Teledyne MX300-I.
However, one of the most important roles played by the user manual is to help in solving problems with Teledyne MX300-I. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Teledyne MX300-I along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center