Instruction/ maintenance manual of the product 9060Z Teledyne
Go to page of 67
TELEDYNE ANALYTICAL INSTRUMENTS Oxygen Transmitter Model 9060Z Technical Manual .
Model 9060Z Manual 2 TABLE OF CONTENTS TABLE OF CONTENTS.............. ...................................... ....................................................... 1 1. IMPORTANT NOTICES.
Model 9060Z Manual 3 5.3 KEYPAD.............. ...................................... ................... ................ ................. ................ 28 5.3.1 Keypad in RUN mode .......
Model 9060Z Manual 4 8.1 FUNCTION SUMMARY TABLE ................ ...................................... ................................ 42 8.2 CALIBRATION MODE FUNCTIONS .............
Model 9060Z Manual 5 Copyright TELEDYNE ANALYTICAL INSTRUMENTS — 2012 This manual describes the transmitter firmware version 1.
Model 9060Z Manual 6 Important Notice Regarding 9060 P.
Model 9060Z Manual 7 1. IMPORTANT NOTICES This manual is intended as a supplement to the 9060Z Operators Manual.
Model 9060Z Manual 8 2. INTRODUCTION 2.1 9060Z TRANSMITTER The Teledyne Analytical Instruments 9060Z oxygen transmitter i.
Model 9060Z Manual 9 Teledyne Analytical Instruments series 9060H oxygen probes and sensors employ state ‐ of ‐ the ‐ art .
Model 9060Z Manual 10 Caution, risk of danger. See additional information in the manual.
Model 9060Z Manual 11 Probe insertion length (from process end of mounting thread to probe sensing tip.
Model 9060Z Manual 12 3.1.4 OXYGEN PROBE MODEL SELECTION GUIDE Heated probes ‐ temperature range 0 ‐ 900°C (1650°F).
Model 9060Z Manual 13 Unheated probes for clean gases ‐ temper atur e range 700 ‐ 1400°C (1290 ‐ 2550°F). *Note: (2) A standard oxygen probe for carburising furnaces, has a 253 MA sheath.
Model 9060Z Manual 14 3.2 OPERATIONAL SPECIFICATIONS Range of outputs: Function Minimum Maximum Linear oxygen #1 0 to 1% 0 to 100% (**) Very low oxygen #1 0 to 0.
Model 9060Z Manual 15 The oxygen deficiency output can be used in the same way as a combustibles analyzer to signal the extent of reducing conditions of combustion processes.
Model 9060Z Manual 16 Probes can be mounted at any angle. However, if the probe is to be mounted on a vertical duct wall, it is better to angle the probe (approx.
Model 9060Z Manual 17 4.3 INSTALLING THE AUXILIARY THERMOCOUPLE Weld a 1/2 inch BSP mounting socket to the flue within about 300 mm (12”), and upstream of the oxygen probe.
Model 9060Z Manual 18 Connection Diagram for 9060Z Transmitter and one or two 9060H Heated Sensors .
Model 9060Z Manual 19 Connection Diagram for 9060Z Transmitter and one or two 9060UL/UH Unheated Probes .
Model 9060Z Manual 20 • The mains power must be either 100/110VAC or 220/240VAC with a mains frequency of 50 .
Model 9060Z Manual 21 Connection of Probe Cable for Unheated Probes Models 9060UL/UH. Connection of Probe Cable for Heated Probes Model 9060H .
Model 9060Z Manual 22 4.8 CONNECTING THE AUXILIARY THER MOCOUPLE (OPTIONAL) For 9060H heated probes, the auxiliary thermo.
Model 9060Z Manual 23 The on ‐ line auto purge and calibration check system is optional. For details on its operation see chapters 4.24, 4.25 and chapter 11.
Model 9060Z Manual 24 network line from the transmitters is taken from the middle of the transmitter network string, a terminating resistor should be enabled with LK3 at each end of the network line.
Model 9060Z Manual 25 high impedance alarm, ‘Probe 1 (2) High Impedance’, is not active. The actual impedance can be displayed on the lower line.
Model 9060Z Manual 26 downwards with the filter removed. Probes can also be mounted horizontally with no filter with some dusts. An occasional automatic back purge is helpful in this case.
Model 9060Z Manual 27 5. DISPLAY AND KEYPAD Operator interaction with the oxygen transmitter is done through a graphical display and 8 keys on the front of the case.
Model 9060Z Manual 28 5.2 OXYGEN DISPLAY UNITS The measured oxygen content displayed on the top line of the .
Model 9060Z Manual 29 SETUP / RUN key Pressing this key will put the transmitter into one of the menu ‐ driven SETUP, C OMMISSIONING or CALIBRATION modes (see chapters 6, 7 and 8 for details).
Model 9060Z Manual 30 FUNCTION up and FUNCTION down keys These two keys allow for the selection of the setup function.
Model 9060Z Manual 31 6. SETUP MODE The SETUP mode is mode is accessed by pressing the SETUP key momentarily so the words ‘Setup Mode’ appear at the bottom of the display.
Model 9060Z Manual 32 6.3 SETUP MODE FUNCTIONS 6.3.1 Probe 1 Offset Each Teledyne Analytical Instruments probe ha.
Model 9060Z Manual 33 NOTE: An asterisk (*) on the end of the line identifies the ite m is enabled by default after a COLD ‐ START. 6.
Model 9060Z Manual 34 7. COMMISSIONING MODE This chapter describes the functions available when the COMMISSIONING mode is entered on the transmitter.
Model 9060Z Manual 35 .
Model 9060Z Manual 36 7.2 COMMISSIONING MODE FUNCTIONS 7.2.1 Internal Date / Time The date and time are used .
Model 9060Z Manual 37 The zero and span of the selected output are set in the following two menus (functions 11&12 and 14&15).
Model 9060Z Manual 38 7.2.8 Flue Pressure Units and Value The 9060Z transmitter is capable of fixed or variable input pressure compensation in the calculation of oxygen from ‐ 1 Atm to +3 Atm.
Model 9060Z Manual 39 7.2.16 Solenoid 1 & 2 Post Freeze Duration of time the transmitter waits before resuming live readings following a gas calibration or purge event.
Model 9060Z Manual 40 All of the items in this list are selected as the default setting.
Model 9060Z Manual 41 7.2.25 Communications Port Options The 9060Z transmitter has a serial communications port available at terminals 18 to 22.
Model 9060Z Manual 42 8. CALIBRATION MODE This chapter describes the functions available when the CALIBRATION mode is selected on the transmitter.
Model 9060Z Manual 43 8.2 CALIBRATION MODE FUNCTIONS 8.2.1 REFERENCE VOLTAGES The calibration of the analogue inputs is based on the voltage of a temper ature compensated voltage reference integrated circuit.
Model 9060Z Manual 44 8.2.8 HEATER SSR FAULT CORRECTION In the event of one of the heater solid state relays .
Model 9060Z Manual 45 9. ALARMS The 9060Z has 4 alarm relays and a built in alarm annunciator. If an alarm occurs, the ALARM light will flash.
Model 9060Z Manual 46 4. ‘Probe 2 High Impedance’ Oxygen probe or electrode failure. The sensor has a high impedance measurement. Replac e the probe.
Model 9060Z Manual 47 19. ‘Gas 1 Calibration Error’ 20. ‘Gas 2 Calibration Error’ At the end of an au.
Model 9060Z Manual 48 level. With heated probes this relay will be de ‐ energised while the probe is heating up from ambient, making the contacts open circuit.
Model 9060Z Manual 49 10. INSTRUMENT CALIBRATION 10.1 CALIBRATION SUMMARY The 9060Z oxygen transmitter has a self calibration and diagnostic system built into the hardware and software.
Model 9060Z Manual 50 11. The output should go 20. 00mA. The value in function #7 (10) can be fine tuned if necessary.
Model 9060Z Manual 51 11. GAS CALIBRATION CHECK AND PURGE The oxygen sensor that is used in the Teledyne Analytical Instruments oxygen probe is extrem ely predictable, stable and reliable.
Model 9060Z Manual 52 12. SOFTWARE UPGRADES The software for the 9060Z series transmitters is contained in memory inside the microprocessor integrated circuit (IC).
Model 9060Z Manual 53 13. TROUBLESHOOTING The Teledyne Analytical Instruments oxygen transmitter has proved an extremely reliable instru ment.
Model 9060Z Manual 54 14. MAINTENANCE 14.1 TRANSMITTER MAINTENANCE The 9060Z transmitter has several hardware checking sy.
Model 9060Z Manual 55 15. INDEX 9060H Oxygen Probes .................................... .................................... ................ ............... ................ 9 Alarm connections .
Model 9060Z Manual 56 Probe offset entry .................. .................. .................. .................. ................. ............. ............ .......... 34 Purge ..............
Model 9060Z Manual 57 APPENDIX 1, CONSTITUENT VALUES FOR VARIOUS FUELS If the transmitter is set up to have r.
Model 9060Z Manual 58 APPENDIX 2, PROBE EMF TABLES ZIRCONIA OXYGEN SENSOR OUTPUT (mV) PROBE TYPE 9060H, SENSOR .
Model 9060Z Manual 59 ZIRCONIA OXYGEN PROBE OUTPUT (mV) PROBE TYPE 9060UL/UH These tables are based on the .
Model 9060Z Manual 60 EMF = ₂ . . Where T is temperature° K Thermocouple .
Model 9060Z Manual 61 APPENDIX 4, MODBUS™ The Teledyne Analytical Instruments 9060Z transmitter has the ability .
Model 9060Z Manual 62 Alarm status is stored as an integer value: 0 clear 2 active 1 .
Model 9060Z Manual 63 .
Model 9060Z Manual 64 .
Model 9060Z Manual 65 9060Z Transmitter Modbus Holding Register Table for firmware v1.16 and newer .
Model 9060Z Manual 66 .
Model 9060Z Manual 67 0x41 Special Instruction Function This command allows you to interact with the transmitter allo wing access to functionality that would otherwise only be accessible via the keypad.
An important point after buying a device Teledyne 9060Z (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Teledyne 9060Z yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Teledyne 9060Z - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Teledyne 9060Z you will learn all the available features of the product, as well as information on its operation. The information that you get Teledyne 9060Z will certainly help you make a decision on the purchase.
If you already are a holder of Teledyne 9060Z, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Teledyne 9060Z.
However, one of the most important roles played by the user manual is to help in solving problems with Teledyne 9060Z. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Teledyne 9060Z along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center