1 319R Introduction 1.0 TELEDYNE BROWN ENGINEERING Analytical Instruments Introduction The Teledyne Brown Engineering Analytical Instruments (TBE) Model 319R Oxygen Detector measures the percent of oxygen in a wide range of gases and gas mixtures.
1.0 Introduction 319R 2 TELEDYNE BROWN ENGINEERING Analytical Instruments 4. Proven Sensor. The 319R uses the Class Z-2 Micro-Fuel Cell, warranted for 6 months, with a life expectancy of 6 months. 5. Analog Voltage Output. The Model 319R provides customer connections to external devices, such as computers, recorders, alarms, etc.
3 319R Installation & Operation 2.0 TELEDYNE BROWN ENGINEERING Analytical Instruments Installation & Operation Precautions 1. Read this chapter in its entirety before installing the unit. 2. Upon receipt of the instrument, inspect the entire unit for damage.
4 2.0 Installation & Operation 319R TELEDYNE BROWN ENGINEERING Analytical Instruments The analysis unit should be installed in the process at a maximum distance of 6 feet away from the main unit, depending upon the level of electromagnetic noise in the operating environment.
5 319R Installation & Operation 2.0 TELEDYNE BROWN ENGINEERING Analytical Instruments Alarm Terminal Connection SPDT relay contacts are provided for each of the 3 alarms. There are 9 terminals available on the back panel of the system for alarm circuit connection.
6 2.0 Installation & Operation 319R TELEDYNE BROWN ENGINEERING Analytical Instruments Setting Alarms The alarms can be adjusted over the full scale range, and they are set as a percent of full scale.
7 319R Installation & Operation 2.0 TELEDYNE BROWN ENGINEERING Analytical Instruments Calibration To calibrate the 319R: 1. Place the power switch in the ON position. 2. Allow the LCD reading to stabilize. 3. Expose the sensor to ambient air (20.9%) or instrument grade air.
8 2.0 Installation & Operation 319R TELEDYNE BROWN ENGINEERING Analytical Instruments.
9 319R Maintenance 3.0 TELEDYNE BROWN ENGINEERING Analytical Instruments Maintenance Replacing the Fuse If the unit will not turn on when plugged in and switched ON (via the power switch), or you suspect a blown fuse, replace the fuse using the following procedure: 1.
3.0 Maintenance 319R 10 TELEDYNE BROWN ENGINEERING Analytical Instruments 5. Replace the fuse holder into its receptacle, pushing in firmly until it clicks. In units with DC power, the fuse is located on the rear panel above the ON/OFF switch. Open the fuse holder by unscrewing and removing the knob marked FUSE.
11 319R Maintenance 3.0 TELEDYNE BROWN ENGINEERING Analytical Instruments Warranty Conditions The Class Z-2 cell used in the 319R is warranted for six (6) months of service from the date of shipment. Customers having warranty claims must return the cell in question to the factory for evaluation.
3.0 Maintenance 319R 12 TELEDYNE BROWN ENGINEERING Analytical Instruments.
13 319R Appendix TELEDYNE BROWN ENGINEERING Analytical Instruments Appendix Specifications Ranges : 0–200 PPM, 0–1999 PPM O 2 , and 0– 25% calibration range Sensitivity: 1% of full scale Signal Output : Standard: 0–10 VDC, negative ground 4-20 mA, negative ground plus contact closures for range identification.
14 Appendix 319R TELEDYNE BROWN ENGINEERING Analytical Instruments Spare Parts List QTY. P/N DESCRIPTION 1 C57304 PC board, main 1 A33748 Thermistor assembly 1 C6689-Z2 Micro-Fuel Cell, Z-2 4 F1189 Fuse, 1A (AC) 4 F51 Fuse, ½A (DC) 1 A59264 Probe to analyzer cable, 6 feet A minimum charge of US $20.
15 319R Appendix TELEDYNE BROWN ENGINEERING Analytical Instruments Drawing List C-57490 Outline diagram A-58154 Probe interconnection diagram D-57302 Electronic PCB schematic diagram C-57305 Power sup.
16 Appendix 319R TELEDYNE BROWN ENGINEERING Analytical Instruments Material Safety Data Sheet Product Name: Micro-Fuel Cells and Super Cells, all classes except A-2C, A-3, and A-5. Electrochemical Oxygen Sensors, all classes except R-19. Mini-Micro-Fuel Cells, all classes.
17 319R Appendix TELEDYNE BROWN ENGINEERING Analytical Instruments Flash Point: na Flammable Limits: na LEL: na UEL: na Extinguishing Media: Use extinguishing media appropriate to surrounding fire conditions.
18 Appendix 319R TELEDYNE BROWN ENGINEERING Analytical Instruments Contact of electrolyte with skin or eyes will cause a burning sensation and/or feel soapy or slippery to touch. Other symptoms of exposure to lead include loss of sleep, loss of appetite, metallic taste and fatigue.
19 319R Appendix TELEDYNE BROWN ENGINEERING Analytical Instruments Section VIII – Control Measures Eye Protection: Chemical splash goggles Hand Protection: Rubber gloves Other Protective Clothing: A.
An important point after buying a device Teledyne 319R (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Teledyne 319R yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Teledyne 319R - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Teledyne 319R you will learn all the available features of the product, as well as information on its operation. The information that you get Teledyne 319R will certainly help you make a decision on the purchase.
If you already are a holder of Teledyne 319R, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Teledyne 319R.
However, one of the most important roles played by the user manual is to help in solving problems with Teledyne 319R. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Teledyne 319R along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center