INSTRUCTION MANUAL FOR TRACE OXYGEN ANALYZER MODEL 311-D RANGES: 0-100 PPM; 0-1000 PPM; 1% AND 100% TELEDYNE ANALYTICAL INSTRUMENTS 16830 CHESTNUT STREET CITY OF INDUSTRY, CALIF. 91749 TELEPHONE: (626) 934-1500 FAX: (626) 961-2538 TWX: (910) 584-1887 TDYANYL COID EASYLINK: 628-31862 Web: www.
TABLE OF CONTENTS MODEL 311D 1. INTRODUCTION 1 1.1 Description 1 1.2 Method of Analysis 1 1.3 Outstanding Features 1 1.3.1 Micro-Fuel Cell 1 1.3.2 Reliable Calibration 1 1.3.3 Integral Power Supply 2 1.3.4 Accuracy and Response 2 1.3.5 Compact Packaging 3 1.
1. INTRODUCTION 1.1 Description : The Teledyne Analytical Instruments (TAI) Model 311-D is a portable trace oxygen analyzer which can be opera ted without an external power source, and reliably calibrated without the use of cumbersome, questionable, “certified” calibration gases.
After the electronics have been properly zeroed (a one-time factory operation), the instrum ent cannot produce an output indication in the absen ce of oxygen; therefore, the need for a “zero” standardization gas is obviated.
Further disassembly may be accom plished by rem oving the backplate assemb ly from its four mounting standoffs, and laying the two separated assem blies out as illustrated on the “Analyzer Wiring Diagram”. The diagram is included among the drawings at the rear of the manual.
necessary or desirable, it must have a metallic diaphragm. Regulators with organic or plastic diaphragms are permeable to oxygen, and if used in the sampling system, will lead to high oxygen readings.
dictated by the customer’s application. If the instrument were being used to certify the oxygen content of a product for delivery, then a calibrati on prior to certification would certainly be in order.
If, on the other hand, the instrument is not to be used imm ediately after calibration, and a low ppm oxygen gas is being employed as a purge, allo w the m anifold to be purged overnight, and then disconnect both male fittings. ALWAYS DISCONNECT THE SOURCE FITTING FIRST, AND IMMEDIATELY THEREAFTER, THE VENT FITTING .
every two (2) days of continuous use. To rech arge the batteries, place the range switch in the “OFF” position and connect the power cord to a convenience outlet.
No tools are required to replace the cell in the in strument. Sim ply unscrew (ccw) the plug at the bottom of the analyzer, and the cell will drop out of the m anifold cavity. Remove the new cell from its package, and carefu lly rem ove the shorting clip.
1. Disconnect cell. 2. Move range switch to “CAL” position. 3. Adjust P1 such that the output of A3, pin 6, measures between 0 and +0.5 mV, ideally +-.3 mV. 4. Adjust P2 for 0 ± 1 mV at output of A2, pin 6. 5. Verify that the offset is the same on all ranges.
OPERATING TEMPERATURE RANGE: 30 ° TO 125 ° F. RESPONSE AND RECOVERY: AT THE SPECIFIED FLOWRATE (0.25 SCFH), THE SENSOR OUTPUT SHOULD BE, IN TEN SECONDS, WITHIN ± 10% OF AN INTRODUCED GAS HAVING A C.
EFFECT OF CO2 ON B-2C CELL LIFE -11-.
RECOMMENDED SPARE PARTS LIST MODEL 311D QTY. P/N DESCRIPTION 5 F-39 MICRO FUSE, 0.25 AMP., FAST-BLOW (FOR 210-240 VAC) 5 F-51 MICRO FUSE, 0.50 AMP., FAST-BLOW (FOR 100-125 VAC) 1 C-6689-A2C MICRO-FUEL.
DRAWING LIST MODEL 311D B-26473 PICTORAL DIAGRAM C-28339 SCHEMATIC C-26606 W IRING DIAGRAM NOTE: The MSDS on this material is available upon request through the Teledyne Environmental Health and Safety Coordinator.
.
Date: 6/20/2006 Time : 01:45:07 PM Title: Indsrsvr01drawingsMASTER26000C26606-14.dwg.
Date: 7/20/04 Time : 10:56:54 AM Title: G:MASTER28000C28339-8.DWG.
An important point after buying a device Teledyne 311-D (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Teledyne 311-D yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Teledyne 311-D - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Teledyne 311-D you will learn all the available features of the product, as well as information on its operation. The information that you get Teledyne 311-D will certainly help you make a decision on the purchase.
If you already are a holder of Teledyne 311-D, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Teledyne 311-D.
However, one of the most important roles played by the user manual is to help in solving problems with Teledyne 311-D. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Teledyne 311-D along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center