Teledyne API, Model 265A Chemiluminescence Ozone Analy zer Manual Addendum, 03389, Rev. B, DCN 5254 HIGHLIGHTS The purpose of this Highlights Page is to list the changes that were incorporated into this Manual per DCN 5254. Chapter / Page Number Description Title Page Changed Revision number and date.
THIS PAGE IS INTENTIONALLY LEFT BLANK.
TELEDYNE INSTRUMENTS Advanc ed Polluti on Instrumentati on A Tel e dyn e T ech nologies C ompany MANUAL ADDENDUM MODEL 265A CHEMILUMINESCENCE OZONE ANALYZER (ADDENDUM TO M200A INSTRUCTION MANUAL #0224.
THIS PAGE IS INTENTIONALLY LEFT BLANK.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 i PRINTED DOCUMENTS ARE UNCONTROLLED SAFETY MESSAGES Your safety and the safety of others is very im portant. We have provided m any important safety messages in this manual.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 ii PRINTED DOCUMENTS ARE UNCONTROLLED TABLE OF CONTENTS SAFETY MESSAGES ................................................................................................
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 iii PRINTED DOCUMENTS ARE UNCONTROLLED LIST OF FIGURES FIGURE 1-1: REMOVAL OF SHIPPING SCREWS & CHECK FOR CORRECT POWER ............ 3 FIGURE 1-2: M265A REAR PANEL .
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 iv PRINTED DOCUMENTS ARE UNCONTROLLED THIS PAGE IS INTENTIONALLY LEFT BLANK.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 1 PRINTED DOCUMENTS ARE UNCONTROLLED 1 GETTING STARTED 1.0 Overview The Model 265A is a close derivative of the Model 200A Chemiluminescence NOx Analyzer.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 2 PRINTED DOCUMENTS ARE UNCONTROLLED 1.2 Electrical and Pneumatic Connections 1. See Figure 1-2 and Figure 1-3 to locate the rear panel electrical and pneumatic connections.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 3 PRINTED DOCUMENTS ARE UNCONTROLLED Figure 1-1: Removal of Shipping Screws & Check for Correct Pow er.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 4 PRINTED DOCUMENTS ARE UNCONTROLLED Figure 1-2: M265A Rear Panel.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 5 PRINTED DOCUMENTS ARE UNCONTROLLED Figure 1-3: Inlet and Exhaust Venting Recommendations.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 Figure 1-4: V/F Board Switch Settings 6 PRINTED DOCUMENTS ARE UNCONTROLLED.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 7 PRINTED DOCUMENTS ARE UNCONTROLLED 1.3 Initial Operation 1. Turn on the instrument power. The display should imm ediately light, showing the computer’s mem ory configuration, then the instrument type - M265A.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 8 PRINTED DOCUMENTS ARE UNCONTROLLED Step 2 - Calibrate the instrument: Zero/Span Calibration Procedure Step Number Action Comment 1. Input zero gas Allow zero gas to enter the sample port on the rear of the instrument.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 9 PRINTED DOCUMENTS ARE UNCONTROLLED Step Number Action Comment 1. Scroll the TEST function menu until the O 3 SLOPE is displayed. The SLOPE value for O 3 should be 1.0 ± 0.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 10 PRINTED DOCUMENTS ARE UNCONTROLLED THIS PAGE IS INTENTIONALLY LEFT BLANK.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 11 PRINTED DOCUMENTS ARE UNCONTROLLED 2 SPECIFICATIONS, WARRANTY 2.1 Specifications Table 2.1: M265A Specifications Ranges In 1 ppb increments from 100 ppb to 20,000 ppb Measurement Mode Single range or AutoRange Measurement Units ppb, µg/m 3 Noise at zero 0.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 12 PRINTED DOCUMENTS ARE UNCONTROLLED 2.2 Warranty ADVANCED POLLUTION INSTRUMENTATION DIVISION (T-API) (02024D) (DCN 4473) Prior to shipme nt, Teledyne API equipm ent is thoroughly inspected and tested.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 13 PRINTED DOCUMENTS ARE UNCONTROLLED 2.3 Test and Calibration Values Table 2-2: Final Test and Calibration Values TEST Values Observed Value Units Nominal Range Reference Section RANGE ppb 10-20000 NOISE ppb 0.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 14 PRINTED DOCUMENTS ARE UNCONTROLLED Table 2-2: Final Test and Calibration Values (Continued) Parameter Observed Value Units Nominal Range Reference Section O 3 Span Conc ppb 0.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 15 PRINTED DOCUMENTS ARE UNCONTROLLED 3 THE M265A OZONE ANALYZER 3.1 Principle of Operation The Teledyne API Model 265A analyzer is designed to measure the concentration of ozone using the chemiluminescence reaction below.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 16 PRINTED DOCUMENTS ARE UNCONTROLLED TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 17 PRINTED DOCUMENTS ARE UNCONTROLLED Figure 3-2 : Pneumatic Diagram.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 18 PRINTED DOCUMENTS ARE UNCONTROLLED THIS PAGE IS INTENTIONALLY LEFT BLANK.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 19 PRINTED DOCUMENTS ARE UNCONTROLLED 4 SOFTWARE FEATURES The M265A software has two major operating modes. The SAMPLE m ode is the normal mode when the instrument is taking data.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 20 PRINTED DOCUMENTS ARE UNCONTROLLED Figure 4-1: Sample Menu Tree.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 Figure 4-2: Setup Menu Tree 21 PRINTED DOCUMENTS ARE UNCONTROLLED.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 22 PRINTED DOCUMENTS ARE UNCONTROLLED THIS PAGE IS INTENTIONALLY LEFT BLANK.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 23 PRINTED DOCUMENTS ARE UNCONTROLLED 5 MAINTENANCE 5.1 Maintenance Schedule NOTE The operations outlined in this chapter are to be performed by qualified maintenance personnel only.
TAPI Model 265A Chemiluminescence Ozone Analyzer Manual Addendum, Rev. B, DCN 5254 24 PRINTED DOCUMENTS ARE UNCONTROLLED THIS PAGE IS INTENTIONALLY LEFT BLANK.
An important point after buying a device Teledyne 265A (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Teledyne 265A yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Teledyne 265A - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Teledyne 265A you will learn all the available features of the product, as well as information on its operation. The information that you get Teledyne 265A will certainly help you make a decision on the purchase.
If you already are a holder of Teledyne 265A, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Teledyne 265A.
However, one of the most important roles played by the user manual is to help in solving problems with Teledyne 265A. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Teledyne 265A along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center