Instruction/ maintenance manual of the product VM700T Tektronix
Go to page of 16
VM700 T (turbo) V ideo Measurement Set Option 11 P AL V ideo Measurements M any capabilities in one instrument – Digitizing waveform monitor – Digitizing vectorscope – Picture Display – Group .
modem to report measurement r esults or control a signal router . A relative to reference mode allows normalizing to a signal source or eliminate signal path errors from the desired measurement.
Vertical interval test signals can be seen very clearly for additional analysis of the signal. These can be printed as support documentation for automatic measurement results. E ven a single horizontal synchronization pulse can be displayed at a high intensity .
Bar Line Time measurement. Chrominance to Luminance Gain and Delay measurement. MEASURE MODE 1,2 BAR LINE TIME A bsolute Relative M easurement Range Mode Mode Accuracy Accuracy B ar Level (b1 or Back Porch) 300 mV to 1.4 V ± 0.5% ± 0.2% S ync Level 50 mV to 600 mV ± 0.
Chrominance Non-Linearity measurement. Color Bar measurement. Luminance Level (mV ) Luminance Level (mV ) Chrominance Level (mV ) Chrominance Level (mV ) Chrominance Phase (deg) Chrominance Phase (deg) Gray Gray Yellow Yellow Cyan Cyan Green Green Magenta Magenta Red Red Blue Blue Black Black 0.
Frequency Response and Group Delay measurement using Sin X/X. Horizontal Timing measurement. M EASURE MODE (continued) FREQUENCY RESPONSE AND GROUP DELA Y A bsolute Relative M easurement Range Mode Mode A ccuracy Accuracy F requency Response to 5 MHz ± 40 dB ± 1.
LINE FREQUENCY M easurement Range Accuracy L ine Frequency ± 3% ± 0.1% F ield Frequency ± 3% ± 0.1% LUMINANCE NON-LINEARITY Absolute Relative M easurement Range Mode Mode A ccuracy Accuracy L uminance Non-Linearity 0 to 100% ± 0.4% ± 0.2% K-factor measurement.
VERTICAL BLANKING M easurement Range Absolute Mode Accuracy E qualizing Pulse Width 80 ns to 1 µs ± 10 ns B road Pulse Width 80 ns to 1 µs ± 10 ns V ertical Blanking Field 1 19 to 30 lines NA Vertical Blanking Field 2 19 to 30 lines NA Noise Spectrum measurement.
AMPLITUDE AND PHASE MEASUREMENTS (continued) M easurement Range Accuracy ITS Element Standard D ifferential Gain 0 to +100% ± 0.3% D2 CCIR Rec. 569 ( Peak and p-p) (0% nominal) D ifferential Phase 0 to 360 deg ± 0.
Unified Unweighted filter response curve per CCIR Recommendation 567. 0 –5 5 08 M Hz d BV –5 – 10 – 15 –2 0 –2 5 –3 0 –3 5 –4 0 – 45 –5 0 1 234 5 67 U nified Luminance weighted filter response curve per CCIR Recommendation 567. 0 – 14 05 MHz dBV .
Measur ement r esults are displayed in an easy-to-read format indicating the time, signal source, measurement, and whether the measured value exceeded caution (*) or alarm (**) limits.
Measurement methods The following paragraphs specify the methods for each Option 11 measurement. Where appropriate, reference is made to the relevant CCIR recommendation. Line blanking timing measurements Color Burst Duration: Measured between the half-amplitude points of the burst chrominance envelope.
Other timing measurements B ar Rise T ime: M easured between the 10% and 90% points on the leading edge of bar . See element B2. A mplitude and phase measurements Sync Amplitude Error: Measured as the difference b etween the sampled sync pulse amplitude and a nominal 300 mV amplitude.
C/L Gain Inequality: Measured as the difference between the sampled peak-to-peak ampli- tude of the 700 mV (nominal) chrominance packet (G1 or G2) and the sampled amplitude of the luminance bar (also nomi- nally 700 mV). Result expressed as a % of sampled bar amplitude.
Frequency response measurements M ultiburst Flag Amplitude: Measured from the center point of the flag top to the ensuing bottom of the flag. Result e xpressed as % of sampled bar a mplitude.
Contact T ektronix: A SEAN / Australasia / Pakistan ( 65) 6356 3900 A ustria + 41 52 675 3777 B alkan, Israel, South Africa and other ISE Countries + 41 52 675 3777 B elgium 0 7 81 60166 B razil &.
An important point after buying a device Tektronix VM700T (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Tektronix VM700T yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Tektronix VM700T - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Tektronix VM700T you will learn all the available features of the product, as well as information on its operation. The information that you get Tektronix VM700T will certainly help you make a decision on the purchase.
If you already are a holder of Tektronix VM700T, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Tektronix VM700T.
However, one of the most important roles played by the user manual is to help in solving problems with Tektronix VM700T. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Tektronix VM700T along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center