100/200 multiranger

Safety Guidelines: Warning notices must be observed to ensure personal safety as well as that of others, and to protect the product and the connected equipment. These warning notices are accompanied by a clarification of the level of caution to be observed.
Qualified Personnel: This device/system may only be set up and operated in conjunction with this manual. Qualified personnel are only authorized to install and operate this equipment in accordance with established safety practices and standards.
Unit Repair and Excluded Liability:
© Siemens Milltronics Process Instruments Inc. 2009









The MultiRanger is available in two models, MultiRanger 100 and MultiRanger 200, and is designed for a variety of applications:
The MultiRanger 100 is a single or dual-point level, one, three, or six relay measurement device. It is equipped with digital communications and offers the latest in echo processing technology and diagnostic features.
The MultiRanger 200 is a single or dual-point, three or six relay device that offers both level and volume measurement. It has Open Channel Monitoring capabilities, a larger number of advanced pump control algorithms, and is equipped with digital communications. It offers the latest in echo processing technology and diagnostic features.
in a residential area may cause interference to several frequency based communications. • Please follow the installation and operating procedures for a quick, trouble-free
installation and to ensure the maximum accuracy and reliability of your MultiRanger.
The manual provides instruction for both MultiRanger 100 and MultiRanger 200 models. For your convenience, the manual uses MultiRanger 100 features as its standard content. Additional MultiRanger 200 features are clearly marked.
The manual is designed to help you get the most out of your MultiRanger, and it provides information on the following:

If you have any questions, comments, or suggestions about the manual contents, please email us at techpubs.smpi@siemens.com.
For the complete library of Siemens Milltronics manuals, go to www.siemens.com/processautomation
Please note their use carefully.
Configuration Examples
The configuration examples used in this manual illustrate the versatility of the MultiRanger. Because there is often a range of ways to approach an application, other configurations may also apply.
In all examples, substitute your own application details. If the examples do not apply to your application, check the applicable parameter reference for the available options.
Should you require more information, please contact your Siemens Milltronics representative. For a complete list of Siemens Milltronics representatives, go to www.siemens.com/processautomation

Location
• 4

Specifications

Control Relays

Weight


with local governing regulations. • This product is susceptible to electrostatic shock. Follow proper grounding
procedures.
All field wiring must have insulation suitable for at least 250 V.
Hazardous voltage present on transducer terminals during operation.
DC terminals shall be supplied from an SELV source in accordance with IEC 1010-1 Annex H.
connections. Use grounding type bushings and jumpers.
Mounting Locations
Avoid

The wall mount and panel mount units install differently. Please follow the specific instructions for your unit.
14.9 mm (0.58")
160.3 mm (6.325") 130 mm (5.125")
91 mm (3.58")
15.2 mm (0.6")
4.3 mm dia. (0.17") four mounting holes
6.6 mm (0.26")
240 mm (9.45") 227 mm (8.93")
Installation
surface for the four screws (customer supplied). 3. Fasten with a long screwdriver.
lid screws (6) · mounting screw holes

motherboard from the enclosure by pulling the board straight out. 3. Drill the required cable entry holes. Make sure conduit holes do not interfere with
the lower areas on the terminal block, circuit board, or SmartLinx card. 4. Attach the conduit to the enclosure using only approved suitable size hubs for
watertight application. 5. Reinstall the motherboard with the mounting screws.
suitable location for conduit entrances
from the signal cables and then wire the cables to the terminal blocks. 3. Tighten the glands to form a good seal.

Installing the panel mount unit requires making a cutout in the panel. The dimensions for the cutout are provided in the illustration below. A full size cutout template is provided with your unit or may be downloaded from www.siemens.com/processautomation

Cutout Instructions

198 mm (7.80")
97 mm (3.82")
36 mm (1.40")
278 mm (10.93")
Once cutout is complete and mounting holes are drilled, follow these steps:
motherboard from the enclosure by pulling the board straight out. 4. Drill the required cable entry holes. Be sure to compensate for panel door
dimensions and make sure conduit holes do not interfere with the lower areas on the terminal block, circuit board, or SmartLinx card. 5. Replace board and fasten the four screws. 6. Place the unit into the panel and insert hexagonal fasteners through bevel slots and
predrilled panel holes. 7. Fasten with wingnuts and hand tighten. 8. Add conduit or glands and wire as required, then replace the lid.


Female Connector for SmartLinx Card
The battery (Rayovac BR2032) has a ten-year life expectancy. Please note the life expectancy may be reduced by ambient temperature. If the unit loses external and battery power, a capacitor will power the RAM for about ten minutes.

align the + and – terminals correctly. 3. Close and secure enclosure lid.

SmartLinx cards are generally pre-installed. If unit does not have a SmartLinx card, follow these steps to install one.

Siemens Milltronics TS-3 Temperature Sensor
Laptop running Dolphin Plus
Customer Alarm, Pump, or Control Device
Customer Device, digital output
Customer Device, analog output
Customer Network or Modem
Display, PLC, Chart recorder, or other Control Device
MultiRanger · Siemens Milltronics Transducer(s) · SmartLinx Card

The terminal board on the MultiRanger allows all inputs and outputs to be connected simultaneously.
TB2
TB1
L2/N L1
TB3
SYNC
12-30 V
SHIELD
RELAY 1
RELAY 3
TS-3
SHIELD
RELAY 4
2 DISCRETE
INPUTS
RELAY 5
COM
RELAY 6
RS485
A
B
Cables
The MultiRanger transceiver requires a shielded two-wire connection to the transducer.
Connection Cable Type mA input and mA output sync, Temperature sensor, discrete input, dc input Transducer
2 copper conductors, twisted, with shield1/drain wire, 300V 0.324 - 0.823 mm2 (22 - 18 AWG) Maximum length: 365 m
Do not use a coaxial transducer cable extension with the MultiRanger. Electrical noise interference affects performance. Relay output AC input
Relay to be copper conductors per local requirements to meet 250 V 5A contact rating.
mA INPUT · DC Version


Run the transducer cable in a grounded metal conduit, separate from other
terminals • Disregard older transducer manuals that recommend these practices
A 0.1 µF (100V or greater) capacitor is included with the MultiRanger for retrofitting old MultiRanger Plus installations. Please see MultiRanger 100/200 Installation (for retrofitting MultiRanger Plus Installations) on page 247.
Relay contacts are shown in the de-energized position. All relays are handled identically and can be configured as positive or negative logic using P118.

Power Failure Relays 1, 2, 4, and 5 are normally open and will fail in the normal state.
Relays 3 and 6 can be wired either normally open or normally closed, and will fail in their de- energized states.

Accurate temperature readings are critical to accurate level measurements because the speed of sound changes, depending on air temperature, and all Siemens Milltronics Echomax and ST-H transducers have an internal temperature sensor.
If the following conditions apply, a separate TS-3 temperature sensor will ensure optimum accuracy:
mA Input [MR 200 only]
For more information, consult the Transducer (P004) and mA Input Parameters (P250, P251, and P252) in the parameter reference section.
For more information, consult the mA output parameters (P200 to P219) in the parameter reference section.

When using multiple ultrasonic level monitors, be sure to run the transducer cables in separate grounded metal conduits.
When separate conduits are not possible, synchronize the level monitors so that no unit transmits while another is waiting for echo reception.
MultiRanger Other Siemens Milltronics Transceiver
Synchronizing with another MultiRanger 100/200, or other Siemens Milltron- ics instruments (DPL+, SPL, XPL+, LU01, LU02, LU10, LUC500, Hydro+, EnviroRanger, MiniRanger):
Before applying power to the MultiRanger for the first time, ensure any connected alarm/control equipment is disabled until satisfactory system operation and performance is verified.
building installation. • A circuit breaker or switch in the building installation, marked as the disconnect
switch, must be in close proximity to the equipment and within easy reach of the operator.

connected to a reliable ground.
L2/N
Digital Communications
Wiring the MultiRanger for communications allows it to be integrated into a full SCADA system or an industrial LAN.
The MultiRanger can also be directly connected to a computer running Dolphin Plus.

RS-485 Connection (see below)

Ground · Connector

Discrete inputs have a positive and negative terminal. Requires an external power supply.
Discrete Input (post.) 1
Discrete Input (pos.) 2
Common (neg.) for Discrete Input


The MultiRanger has two modes of operation: RUN and PROGRAM.
In RUN mode, the MultiRanger detects material level and provides control functions. The MultiRanger automatically starts in RUN mode when power is applied.
System status is shown on the unit’s LCD, or on a remote communications terminal.


RUN Mode PROGRAM Mode 1 index type (see below) index type (see next table) 2 index index 3 primary reading parameter value 4 units units 5 hi and hi hi alarm designation auxiliary function 6 level display n/a 7 filling display scroll access tag 8 emptying display scroll access tag 9 lo and lo lo alarm designation n/a 10 relay # programmed relay # programmed 11 relay # activated relay # activated 12 auxiliary reading parameter number 13 normal operation: n/a 14 failsafe operation: n/a


Readings in RUN Mode
Change the displayed values with the keys on the hand programmer. All readings are shown in the Auxiliary field, except for the totalizer and P920.
P920
Accumulated pump running hours2 for numbered pump.
Hold number key for five seconds to display the number of accumulated pump starts2 for numbered pump.
[MR 200 only] Eight-digit totalizer, uses index and reading areas, press again to toggle, P737 sets default. Used for OCM and Pumped Volume.
[MR200 only] Head measurement P926
[MR 200 only] Instantaneous flow based on head (OCM) P925
mA Output value P203
Rate of level change P707
Failsafe Time Left (in %). When the Reading is updated, this value (Auxiliary Reading) resets to 100 and begins to decrease until the next valid measurement is made. If the Failsafe Time Left reaches 0, LOE flashes in the display. Hold for four seconds to show echo confidence P805
+ ### Display the value of the entered parameter which is global or indexed by transducer Auxiliary reading displays parameter specified in P731 P731
P

Status parameters give the operating status of the MultiRanger. You can access parameters with the hand programmer (see page 32) or with Dolphin Plus (see page 27). Remote SCADA system access is also possible.
Parameter Values P203 mA Output Value 0 to 22 – Current mA output P254 [MR 200 only]
Scaled mA Input Value 0 to 9999 – Current mA input after scaling
P275 Scaled Discrete Input Value Shows current value of discrete input,
values vary by DI function P322 [MR 200 only]
LCD Total Low The last four digits of the totalizer
P323 [MR 200 only]
LCD Total High The first four digits of the totalizer
P341 Run Time The number of days the MultiRanger has been operating P342 Start Ups The number of times power has been cycled P664 Temperature Current temperature measured by transducer P707 Rate Value Current rate of material level change P708 [MR 200 only]
Volume Rate Display Current rate of material volume change
P729 Scan Time Seconds since last level scan P806 Echo Strength Strength of primary echo P920 Reading Measurement Current primary reading P921 Material Measurement Current level from P007–Span P922 Space Measurement Empty space above the material level P924 [MR 200 only]
Volume Measurement Current volume value, if programmed
P925 [MR 200 only]
Flow Measurement (OCM) Current flow value, if OCM programmed
P926 [MR 200 only]
Head Measurement (OCM) Current level, if OCM programmed
P927 Distance Measurement Distance from transducer face to material

RUN mode provides numerous parameters and variables that you can track on the display (see Display on page 22).
The LCD displays EEEE if a value is too long.
Parameter Action P060 Decimal Position Sets maximum number of decimals P061 Convert Reading Scales the reading to fit P062 Offset Reading Shifts the reading up or down by a fixed amount
To reference the displayed level to sea level, enter the distance in Units (P005), between Empty (P006) and sea level. (Enter a negative value if Empty is below sea level.)

P062 is the distance between sea level and Empty.
The Auxiliary Reading area of the LCD displays parameter values while leaving the primary reading on screen.


To maintain a constant variable display in the auxiliary reading area, set the default.
Example:
To leave the level reading on the screen and view the echo confidence in the auxiliary reading field, set the following parameter:
P730 G 805 Auxiliary field defaults to P805
To display a second auxiliary reading, press in RUN mode.
Parameter Index Value Description
P731 G 912 Shows P912–Transducer Temperature
During differential or average operation (P001 = 4 /5), the display scrolls sequentially through Point Numbers 1, 2, and 3. Point 3 is the difference between (or average of) Points 1 and 2.
Parameter Index Value Description
P732 G 5 Hold each value for 5 seconds
See Parameter Indexing on page 37. All the instructions in the following procedures apply to the hand programmer and assume that the MultiRanger is activated.

The MultiRanger is programmed by setting its parameters to match your specific application. Most parameters are indexed, allowing you to set the parameter to specific conditions and to more than one input or output. When the MultiRanger is in PROGRAM mode, you can change these parameter values and set operating conditions.
Please refer to the Parameter Reference section on page 117 for a full listing and explanations of parameter values.
The MultiRanger’s primary programming is by the hand programmer. Other access is available through Dolphin Plus software (purchased separately).


The hand programmer gives you direct access to the MultiRanger.
For your convenience, the programmer has a magnetic mounting strip on the back. Keep programmer nearby for easy access.
Point the programmer at the IR port above the display and press the keys.
Unless otherwise noted, each valid key press should produce a change in the LCD. Verify when programming the unit.


Keys Programming Mode Run Mode 1 8-digit Totalizer (toggle) [MR 200 only] 2 Pump Running Time
3 Head [MR 200 only] 4 Flow based on Head [MR 200] 5 mA Output
Units or % Units or % (Program Mode) (Key 2) Next Display Field Pause Display Toggle

Operation
Use Dolphin Plus software to configure, monitor, tune, and diagnose the MultiRanger from a PC or directly in the field with a laptop.
Dolphin Plus is easy to install and easy to use. Just load the software from the CD onto a desktop PC or Laptop and then set up or modify complete parameter configurations in a Windows®2environment.
After configuration, you can edit parameters, upload and download parameter sets to and from disk, and use parameter sets saved from other instruments. You can also work with echo profiles for fine tuning without the need for special instruments. Built-in Quick Start features and Help files guide you through the entire process.


The toolbar buttons provide quick access to Dolphin Plus features.

open the vertical echo profile and tank mimic window

Operation
SIMATIC PDM is a software package for parameterizing, commissioning, diagnosing and maintaining process devices. For the MultiRanger 100/200, SIMATIC PDM connects directly to the device using Modbus over Port 1 or Port 2. The MultiRanger 100/200 comes with Port 1 set for communications to SIMATIC PDM.
SIMATIC PDM contains a simple process monitor of the process values, alarms and status signals of the device. Using SIMATIC PDM you can
process device data.
More information about SIMATIC PDM is available at www.siemens.com/processinstrumentation go to Products & Solutions > Products & Systems > Communication and Software > Process Device Manager. Please consult the operating instructions or online help for details on using SIMATIC PDM. An Application Guide on using MultiRanger 100/200 with PDM and Modbus is available on our website: www.siemens.com/processautomation
To use Process Device Manager (PDM) with MultiRanger 100/200, you need the Device Description for MultiRanger 100/200, which will be included with new versions of PDM. You can locate the Device Description in Device Catalog, under Sensors/Level/ Echo/ Siemens Milltronics. If you do not see MultiRanger 100/200 under Siemens Milltronics, you can download it from our website: www.siemens.com/processautomation Go to the MultiRanger product page and click Downloads. After downloading the DD file, you need to execute DeviceInstall.

All the instructions in the following procedures apply to the hand programmer and assume that the MultiRanger is activated.


value is shown. 4. Enter the new value, and press ENTER . The MultiRanger interprets the value, either accepting or replacing it with a valid value.

The Lock parameter P000 secures the MultiRanger against parameter changes via the handheld programmer. The unit can still be put into PROGRAM mode when locked, and parameter values can be viewed, but no parameter values can be changed.
When P000 is set to 1954, programming is enabled. To disable programming, enter another value.
P000 (1954) is a fixed value password. Therefore, you should use other means to secure the MultiRanger if security is a concern.
P000 Lock also controls how simulations affect control relays. By default, control relays are unaffected by simulation levels. But if P000 is set to –1, they react to the simulated level. See Parameters P925–P927 on page 220 for running a simulation.
Many parameters can be viewed either in measurement units (P005) or as a percentage. View the parameter and then press MODE to toggle between units and percentage. The LCD shows the selected measurement type, either units (m, ft) or percentage (%).
Percentage is also available when showing flow and volume with 100%, based on the parameter that defines the maximum.
Measurement Maximum Volume P051 Flow P604
Parameter values indicating status only. They cannot be altered.
Parameter values common to all inputs and outputs on the MultiRanger.
When a global parameter is accessed, the index display automatically disappears. When a non-global parameter is accessed, the index display reappears showing the last index number.

Parameter default values are indicated with an * in the parameter tables.
1954 * OFF: programming permitted -1 Simulation Controls (relays energize based on simulated level) other ON: Lock activated and programming not permitted via the handheld programmer
The asterix identifies 1954 as the default value.
Returning a parameter to factory default.
Returns all parameters to original values.
Use Conditions:
When the dual point option is enabled, P999 is indexed by transducer. Use index 00 to reset the entire MultiRanger.
The following readouts are shown when the MultiRanger cannot display a number.
Display Definition Parameter has not been set
All values not same when viewing index 0
Value too large for four-digit display
P000 Lock · Primary Index Global · Value

The other method for changing parameter values is with Dolphin Plus software. It lets you access the MultiRanger from a PC or on site with a laptop and change MultiRanger parameters.
Most examples in this manual use the icons from the hand programmer but nearly all functions are also available through Dolphin Plus.


Parameters are indexed when they apply to more than one input or output. The index value defines the input/output for that parameter. Indexed parameters contain a value for each index, even if that index is not used.
MultiRanger Display
The index number and the index values are displayed above the parameter indicator on the LCD.
Index Value Index Number


Primary Index: relates to direct input or output and can refer to relays, communications ports, and other parameters. In parameters that allow secondary indexes, the primary index is often referred to as a point.
Secondary Index: relates to previously indexed parameters where the parameter requires a second index, permitting multiple values on an indexed input or output.
3 4 5 6 2 1
Parameters with a secondary index permit multiple values for a primary index (point). For example, a volume calculation based on vessel characterization breakpoints requires a distinct set of breakpoints for each measured point.
Thus the primary index refers to the measurement point, and each secondary index refers to a characterization breakpoint value.

Example [MR 200 only] P054 provides up to 32 breakpoint levels used with P055 (Volume Breakpoint) for universal volume calculation. The illustration indicates how you can set secondary indexes to specific functions.

A B C P054 [1,6] = 2m P054 [1,3] = 1.5m P054 [1,1] = .75m P054 [2,6] = 8m P054 [2,3] = 5.5m P054 [2,1] = 2.75m
The MultiRanger startup varies between single and dual point models.

Note: The number of points is an order option and is set by the factory. On a single point model, the index for P001 is global (G). On a dual point model the index for P001 is 1 or 2. To check the model in use, enter P001 on handheld programmer. For a single point device, parameter P001 will not show an entry for index number on the LCD. (See MultiRanger Display on page 36 for location of index number on the LCD.) Parameter P001 will display an index number for a dual point model.
The MultiRanger starts in DISTANCE mode with the transducer preset for the XPS-10 and an empty distance of 5 meters. Change the following parameters to reflect your application parameters.

Parameter Index Value Description P001 G 1 Operation = level P002 G 1 Material = liquid P003 G 2 Maximum Process Speed = medium P004 G 104 Transducer = XPS-15 P005 G 1 Units = meters P006 G 12 Empty = 12m P007 G 10 Span = 10m

For differential or average operation with a single-point MultiRanger, set P001 to 4 (differential) or 5 (average) and connect two transducers of the same type. All of the relevant parameters then become indexed by the correct transducer:
Ring indicates relay is configured
Dot indicates relay is energized/de- energized
Index Description 2 indexed by Transducer One or Two 3 indexed by level measurement 1 = Transducer One 2 = Transducer Two 3 = Calculated Level (average or difference)
The MultiRanger starts in an OFF state and does not take level measurements. For measurement setup, configure these basic parameters:
If the application uses two measurement points, provide the basic information for each measurement point separately:
P001 1 1 Operation = level
2 3 Operation = distance
P002 1 1 Material = Liquid 2 1
P003 1 2 Max. Process Speed = medium 2 3 Max. Process Speed = fast
P004 1 104 Transducer = XPS - 15 2 102 Transducer = XPS - 10 P005 G 1 Units = meters
P006 1 12 Empty = 12m 2 4 Empty = 4m
P007 1 11 Span = 11m 2 3.5 Span = 3.5m
Relay Number

For differential or average operation dual-point MultiRanger, set P001 to 4 (differential) or 5 (average) and connect two transducers of the same type.
All the relevant parameters are then indexed by the correct number:
Index Description 2 indexed by Transducer One or Two 3 indexed by level measurement 1 = Transducer One 2 = Transducer Two 3 = Calculated Level (average or difference)
The following information will help you configure your MultiRanger for optimal performance and reliability.
The response rate of the device influences the measurement reliability. Use the slowest rate possible with the application requirements.
The response rate is also important to functions connected to the filling or emptying indicators.
The dimensions of the vessel, wet well, or reservoir (except empty and span) are only important if you require volume.
Volume is required to report the level value in terms of volume. The pumped volume function can also report pumped volume or pump efficiencies.
The failsafe parameters ensure that the devices controlled by the MultiRanger default to an appropriate state when a valid level reading is not available.
![Page 51 of the Siemens 200 user manual — Rate of Change [MR 200]](/cdn/reader/82/829c15aaebfbf17c854075ac1e0dbc01f9b82abbfd8c8576c5c8e04c8c601513/images/page-51.webp?w=400)
Relays are the primary controls of external devices such as pumps or alarms.
The MultiRanger comes with extensive control and alarm functions.
Depending on the model, up to six onboard multi-purpose relays are provided on the MultiRanger. Each relay may be independently assigned to one function and has a corresponding status icon on the LCD.
The relay functions fall under three modes of operation:
Mode Function alarm alarm ON = LCD Icon ON = relay coil de-energized pump pump ON = LCD Icon ON = relay coil energized miscellaneous contact closed = LCD Icon ON = relay coil energized
Relay Function
Note: The MultiRanger 100 or 200 can be programmed with relays. The number of relays installed depends on the model. To determine the number of available relays that can be utilized in your MultiRanger 100 or 200, open the lid and count the large white relays to the left of the display. It is important to count the number of on-board relays as the software will allow programming of up to six relays whether they are installed or not.
In high alarm, the function goes on when the level rises to the ON setpoint and goes off when the level lowers to the OFF setpoint. In low alarm, the function goes on when the level lowers to the ON setpoint and goes off when the level rises to the OFF setpoint.
The relay will be in alarm if the level is inside the zone between the setpoints.
The relay will be in alarm if the level is outside the zone between the setpoints.
In filling alarm, the function goes on when the rate of filling increases to the ON setpoint and goes off when the rate of filling drops to the OFF setpoint. In emptying alarm, the function goes on when the rate of emptying increases to the ON setpoint and goes OFF when the rate of emptying drops to the OFF setpoint. For emptying alarm, the setpoints must be entered as negative values.

In high alarm, the function goes on when the temperature rises to the ON setpoint and goes off when the temperature lowers to the OFF setpoint. In low alarm, the function goes on when the temperature lowers to the ON setpoint and goes off when the temperature rises to the OFF setpoint.
The function goes on when the fail-safe timer expires. The function goes OFF when a valid echo is received (fail-safe timer is reset).
Relays
In pump down, the function goes on when the level rises to the ON setpoint and goes off when the level lowers to the OFF setpoint. In pump up, the function goes on when the level lowers to the ON setpoint and goes off when the level rises to the OFF setpoint.
Refer to Totalizing Pumped Volume on page 65. Relays are normally de-energized, contact closure is approximately 200 mSec duration.
If the ON setpoint is higher than the OFF setpoint, the relay operates as:
If the ON setpoint is lower than the OFF setpoint, the relay operates as:
The ON and OFF setpoints can not be the same on an individual relay but may be common to other relays. The dead band or hysteresis is the difference between the ON and OFF setpoints. For in and out of bounds level alarms, the hysteresis is set at ± 2 % of span from either boundary.

When the fail-safe timer expires, pump control relays respond as previously described. However, alarm relays will respond in the following manner:
Fail-Safe Mode Relay Status High Alarm Low Alarm Fail-Safe High ON OFF Fail-Safe Low OFF ON Fail- Safe Hold HOLD HOLD
Upon entering the program mode, all pump control relays will be turned OFF. Alarm relays will hold their prior status.
Relay States
The relays on the MultiRanger are completely programmable, allowing for any control scheme.
Relay Types Relay 1,2,4,5 – NO (Form A) Relay 3,6 – NO / NC (Form C)
Some parameters affect how relays react during normal conditions:
Sets the MultiRanger to a preset application. These preset applications quickly set up the MultiRanger with a minimum number of parameters.
Sets the default state differently, depending on whether the relay is programmed as an alarm or a control.
The alarm function de-energizes the relay coils. During normal operation (no alarms), the relay coils are energized.

The control function energizes the relay coils. When the instrument is at rest (no controls operating) the relay coils are de- energized.
Sets the process point at which the relay is tripped.
Sets the process point at which the relay is reset.
Affects relay reaction. Reverses the logic (normally-open to normally-closed or vice versa).
P129–Relay Failsafe
Changes how individual relays react to a failsafe condition on the instrument.
Checks the application wiring by forcing a relay control function, such as a level alarm or pump control setpoint. Ensure all the relay programming and wiring works properly.
Please verify that ON and OFF respond correctly. Use P119 as a final test once all of the relay programming is done.
The flexibility of the relay functions ensures that the MultiRanger can support relay wiring for different systems and applications. Use the following as a guide to the most common parameters.
[MR 100]: When a setpoint is reached, the corresponding action is taken. The setpoint can be an ON or OFF setpoint related to a process variable.
[MR 200]: The setpoint can be an ON or OFF setpoint related to a process variable, or a timed setpoint based on interval and duration.

[MR 100]: Functions affected by setpoint are configured by parameters that determine the application requirements such as timing. P111 Pump and Control functions (see page 134) sets the functions requirements.
[MR 200]: Functions affected by setpoint are configured by parameters that determine the application requirements such as timing. P111 Pump and Control functions (see page 134 sets the function requirements. Other function parameters:
Normal operating conditions means that alarms are off and pumps are on. This can be reversed using P118–Relay Output Logic.
Relay Failsafe
P129–Relay Failsafe
Adjusts how individual relays react to a failsafe condition. Relays can be set to:

Preset applications set up the relay parameters to predetermined values shown below:
Value # Parameters Affected Off 0 All relays set to OFF Wet Well 1
Pump down with the following settings:
Parameter Relay # 1 2 3 4 5 6 P111 52 52 1(H) 1(L) 0 0 P112 70% 80% 90% 10% – – P113 20% 20% 85% 15% – – Wet Well 2
Pump down with the following level and rate settings:
Parameter Relay # 1 2 3 4 5 6 P111 52 52 1(H) 1(L) 0 0 P112 70% 80% 90% 10% – – P113 20% 20% 85% 15% – – P121 1 P121 sets the pump relays to accept control by rate of level change once the first ON setpoint is reached. Because the pumps are started by rate, you must change P703 to desired empty rate. Reservoir 1
Pump up with the following level settings:
Parameter Relay # 1 2 3 4 5 6 P111 52 52 1(H) 1(L) 0 0 P112 30% 20% 90% 10% – – P113 80% 80% 85% 15% – – Reservoir 2
Pump up with the following level and rate settings:
Parameter Relay # 1 2 3 4 5 6 P111 52 52 1(H) 1(L) 0 0 P112 20% 20% 90% 10% – – P113 80% 80% 85% 15% – – P121 1 P121 sets the pump relays to accept control by rate of level change once the first ON setpoint is reached. Because the pumps are started by rate you must change P702 to desired fill rate. Screen
Differential control of a screen or rake:
Parameter Relay # 1 2 3 4 5 6 P110 3 1 2 3 0 0 P111 50 1(H) 1(L) 1(H) – – P112 80% 90% 10% 90% – – P113 20% 85% 15% 10% – – Alarms
General alarms at four setpoints:
Parameter Relay # 1 2 3 4 5 6 P111 1(H) 1(L) 1(HH) 1(LL) 0 0 P112 80% 20% 90% 10% – – P113 75% 25% 85% 15% – –

Backup level override provides the option of overriding the ultrasonic input with another contacting point level device, for example, the Pointek CLS200. The ultrasonic reading is fixed at the programmed switch level until the discrete input is released and the ultrasonic device makes its decisions based on the override value.
Sets the discrete input as the source of a level reading override.
Substitutes value for current reading when the discrete input (P064) is enabled. Value is added in current units and is valid only for the following:
Example:
A high level backup switch is connected to Digital Input Two in the same application as Transducer One at level value 4.3 m.
Parameter Index Value P064 1 2 P064 2 0 P065 1 4.3 P065 2 –
When the level rises to 4.3 m and the switch is activated, the reading is forced to 4.3 m where it stays until the switch is de-activated.
Sets the time (in seconds) used to calm the override condition input.
Settings

Wiring the Discrete Inputs
Normal state is standard operation, with the MultiRanger sensing the material level and controlling the pumps.
The discrete input contacts are either normally-open or normally-closed when the system state is normal.
Example:
Normal state for a backup high level switch is open, and the contacts on the discrete input are wired as normally-open.
See Discrete Inputs on page 21 for complete details on wiring the discrete inputs. To override a level using a discrete input, see Backup Level Override on page 47.
The P270 series of parameters permits control over the discrete input.
DI State P270 Setting Normally Open Normally Closed
The current value of the discrete input is reported in P275:
P275 Setting MultiRanger State 0 1 Normal State Exception State

To integrate the MultiRanger with other equipment, use the mA input and outputs.
The mA input can be used as a level measurement or can be passed on to a SCADA system.
Parameter Index Value Description P004 1 250 Transducer = mA input 1 P250 1 2 Scale = 4 to 20 mA P251 1 0 4 mA = 0% of span P252 1 100 20 mA = 100% of span P253 1 0 Do not damp the input signal
To pass the mA input on to a SCADA system, read the value from the appropriate communication registers. For more information, go to the MultiRanger Communications section on page 89.
The MultiRanger has two mA outputs, used to send measurements to other devices.
Configuring the mA output to send a 4 to 20 mA signal scaled from 10% to 90% of span of the second transducer:
Parameter Index Value Description P200 1 2 set to 4 to 20 range P201 1 1 send mA proportional to level reading P202 1 2 base mA on level point 2 P210 1 10 set 4 mA at 10% of span1
P211 1 90 set 20 mA at 90% of span2
P219 1 0 set failsafe action as 0 mA

a.Attach ammeter to MultiRanger mA output. b.Access P214, Index 1 (for mA output 1) or 2 (for mA output 2). Press CLEAR and
ENTER . The ammeter should show a value near 4 mA. c.Enter the exact value displayed on the ammeter into P214 (Index 1 or 2). d.The ammeter should then read exactly 4.00 mA. The unit is now calibrated for 4 mA for the receiving device.
a.Attach ammeter to MultiRanger mA output. b.Access P215, Index 1 (for mA output 1) or 2 (for mA output 2). Press CLEAR and
ENTER . The ammeter should show a value near 20 mA. c.Enter the exact value displayed on the ammeter into P215 (Index 1 or 2). d.The ammeter should then read exactly 20.00 mA. The unit is now calibrated for 20 mA for the receiving device.
Checks that the external device can track the entire 4 to 20 mA range sent by the MultiRanger.
the simulation.

![Siemens 200 — Volume [MR 200] diagram (page 61)](/cdn/reader/82/829c15aaebfbf17c854075ac1e0dbc01f9b82abbfd8c8576c5c8e04c8c601513/images/fig-61-1.webp)
Volume is a feature of the MultiRanger 200 only. Volume is used in two situations:
If you require this functionality, please contact your local Siemens Milltronics representative at www.siemens.com/processautomation.
When using volume, readings are given in arbitrary units specified in P051.
The default is 100, which gives a reading in percent of total. Use whatever units you want here. If the value is too large for the four-digit LCD, use a larger unit.
If a wet well has a maximum capacity of 250,000 liters, use the value 250.0 for P051 and set the reading in 1000s of liters.
There are many common tank shapes to select from. (See P050. If possible, use one of these.) Each tank shape uses the Empty distance (P006) in its calculations of volume.
Some tank shapes also require extra dimensions to calculate the volumes. Do not estimate these values. They must be correct to ensure the accuracy of your volume calculations.

To configure volume for a tank with a half-sphere bottom, set the following:
Parameter Index Value Description P050 1 4 selects the correct tank shape P051 1 100 sets maximum volume at 100 (percent) P052 1 1.3 sets A to 1.3m

If you cannot use a pre-defined tank, then use one of the universal tank shapes and program the characterization curve.
as steps in the well wall).
VOLUME (P055)
FLOW
LEVEL (P054) 0,0

Volume
2 0.8 3 2.0 4 3.5 5 4.1 6 4.7 7 5.1 8 5.2 9 5.3 10 5.4 11 5.5 12 5.6 13 6.0 14 7.2 15 9.0
Determines the Level breakpoints at which the volumes are known.
2 2.1 3 4.0 4 5.6 5 5.9 6 6.3 7 6.7 8 7.1 9 7.8 10 8.2 11 8.8 12 9.2 13 10.9 14 13.0 15 15.0
Determines the volumes which correspond to the level breakpoints. The universal calculations interpret between the breakpoints to produce an accurate model of the volume at all level readings.
Settings P050 = 9 for linear approximation P050 = 10 for curved approximation
Linear approximation uses a linear algorithm; curved approximation uses a cubic spline algorithm.

Level
The level alarm is the most common. Use this alarm to warn you when your process is in danger of being upset due to high or low levels.
Generally, the four alarms used are Hi, Hi Hi, Lo, and Lo Lo.
Alarms
P006
P007
Prerequisite: You must know the details of your application and substitute the values for the sample values provided. If you are bench testing the unit, then set your test values to be the same as the sample values.
Value Description P001 G 1 Operation = level P002 G 1 Material = liquid P003 G 2 Maximum Process Speed = medium P004 G 102 Transducer = XPS-10 P005 G 1 Units = meters P006 G 1.8 Empty = 1.8m P007 G 1.4 Span = 1.4m

To set Relay Five to a standard level alarm (Hi Hi, Hi, Lo, Lo Lo) do the following:
Available designations:
Alarm Designation Hi Hi HH Hi H Lo L Lo Lo LL
Rate alarms can trigger an alarm if the vessel is filling/emptying too quickly.
Parameter Index Value Description P111 5 4 These settings trip the alarm when the reservoir is filling faster than 1 m per minute and reset it at 0.9m per minute. P112 5 1m P113 5 0.9m
Parameter Index Value Description P111 5 4 These settings trip the alarm when the reservoir is emptying faster than 10% of span per minute and reset the alarm when emptying falls to 5%.
![Page 66 of the Siemens 200 user manual — Temperature [MR 200]](/cdn/reader/82/829c15aaebfbf17c854075ac1e0dbc01f9b82abbfd8c8576c5c8e04c8c601513/images/page-66.webp?w=400)
Use the bounded range alarms to detect when the level is inside or outside of the range. By using a bounded range alarm, you can effectively put two level alarms (high and low) on one relay.
Parameter Index Value P111 5 3 P112 5 1.3 P113 5 0.3 P116 5 0.05
Results:
Parameter Index Value P111 5 2 P112 5 1.3 P113 5 0.3 P116 5 0.05
Results:
Activates an alarm if transducer cable circuit enters a shorted or opened state.
Parameter Index Value Description P111 5 7 Alarm on transducer cable fault P110 5 1 Alarm on Transducer One
Use the temperature alarm to activate an alarm when the temperature reaches the ON setpoint (P112). This alarm uses the same setpoint parameters as the level alarms (P112 and P113).

With P112 and P113, you can set a high alarm (P112 > P113) or a low alarm (P112 < P113).
Parameter Index Value Description P111 5 5 Alarm on temperature P112 5 45 ON setpoint at 45 °C P113 5 43 OFF setpoint at 43 °C P110 5 1 Take the temperature reading from Transducer One
The temperature source can be the temperature sensor built into the transducer or an external TS-3, as set by P660.
Loss of Echo (LOE)
Parameter Index Value Description P110 5 1 Alarm on LOE for Transducer One P111 5 6 Alarm on LOE P070 G 0.5 Trip alarm when 0.5 minutes (30 seconds) pass without detecting a valid echo.

Setting a Pump Down Group
Setting a group of three pumps to pump down a wet well.

Set the Common Parameters
Prerequisite: Substitute the details of your application in place of the sample values provided. If you are bench testing the unit, set your test values to be the same as the sample values.

Value Description P001 G 1 Operation = level P002 G 1 Material = liquid P003 G 2 Maximum Process Speed = medium P004 G 102 Transducer = XPS-10 P005 G 1 Units = meters P006 G 1.8 Empty = 1.8m P007 G 1.4 Span = 1.4m


Parameter Index Value Description P111 1 52 Sets the pump relays (index 1, 2, and 3) to ALTERNATE DUTY ASSIST. P111 2 52 P111 3 52
Parameter Index Value Description P112 1 1.0m Sets the three setpoints for the pump relays. The first cycle will use these setpoints. Subsequent cycles rotate the setpoints among the pumps.
P112 2 1.1m P112 3 1.2m
Parameter Index Value Description P113 0 0.5m By using index 0 all six relays are set at the same time, including any alarm relays. Use with caution.
Sets a group of three pumps to pump up a reservoir.


Prerequisite: Substitute the details of your application in place of the sample values provided. If you are bench testing the unit, set your test values to be the same as the sample values.
Value Description P001 G 1 Operation = level P002 G 1 Material = liquid P003 G 2 Maximum Process Speed = medium P004 G 102 Transducer = XPS-10 P005 G 1 Units = meters P006 G 1.8 Empty = 1.8m P007 G 1.4 Span = 1.4m
Parameter Index Value Description P111 1 52 Sets the pump relays (index 1, 2, and 3) to ALTERNATE DUTY ASSIST. P111 2 52 P111 3 52
Set the Relay ON Setpoints
Parameter Index Value Description P112 1 0.4m Sets the three setpoints for the pump relays. The first cycle will use these setpoints. Subsequent cycles rotate the setpoints among the pumps.
P112 2 0.3m P112 3 0.2m
Parameter Index Value Description P113 0 1.3m By using index 0 all six relays are set at the same time, including any alarm relays. Use with caution.
For more information, see Appendix D: Pump Control Reference on page 239.

Set Relays to ALTERNATE DUTY BACKUP [MR 200]
Parameter Index Value Description P111 1 53 Sets the pump relays (index 1, 2, and 3) to ALTERNATE DUTY BACKUP. P111 2 53 P111 3 53
Parameter Index Value Description P112 1 0.4m Sets the three setpoints for the pump relays. The first cycle will use these setpoints. Subsequent cycles rotate the setpoints among the pumps.
P112 2 0.3m P112 3 0.2m
Parameter Index Value Description P113 0 1.3m By using index 0 all six relays are set at the same time, including any alarm relays. Use with caution.
Set Relays to FIXED DUTY ASSIST
Parameter Index Value Description P111 1 50 Sets the pump relays (index 1, 2, and 3) to FIXED DUTY ASSIST. Multiple pumps can run simultaneously. P111 2 50 P111 3 50
Parameter Index Value Description P112 1 0.4m Sets the three setpoints for the pump relays. The setpoints remain attached to the pump relays. P112 2 0.3m P112 3 0.2m
Parameter Index Value Description P113 0 1.3m By using index 0 all six relays are set at the same time, including any alarm relays. Use with caution.
![Page 72 of the Siemens 200 user manual — Set the Relay OFF Setpoints [MR 200]](/cdn/reader/82/829c15aaebfbf17c854075ac1e0dbc01f9b82abbfd8c8576c5c8e04c8c601513/images/page-72.webp?w=400)
Parameter Index Value Description P111 1 51 Sets the pump relays (index 1, 2, and 3) to FIXED DUTY BACKUP. Only one pump will ever run at one time. P111 2 51 P111 3 51
Parameter Index Value Description P112 1 0.4m Sets the three setpoints for the pump relays. The setpoints remain attached to the pump relays. P112 2 0.3m P112 3 0.2m
Parameter Index Value Description P113 0 1.3m By using index 0 all six relays are set at the same time, including any alarm relays. Use with caution.
Parameter Index Value Description P111 1 54 Sets the pump relays (index 1, 2, and 3) to SERVICE RATIO DUTY ASSIST P111 2 54 P111 3 54 P122 1 25 Sets the ratio to: 25% – Pump One
50% – Pump Two 25% – Pump Three P122 2 50 P122 3 25
Parameter Index Value Description P112 1 0.4m Sets the three setpoints for the pump relays. The first cycle will use these setpoints. Subsequent cycles rotate the setpoints among the pumps.
P112 2 0.3m P112 3 0.2m
Parameter Index Value Description P113 0 1.3m By using index 0 all six relays are set at the same time, including any alarm relays. Use with caution.

Parameter Index Value Description P111 1 56 Sets the pump relays (index 1, 2, and 3) to FIFO DUTY ASSIST. P111 2 56 P111 3 56
Parameter Index Value Description P112 1 0.4m Sets the three setpoints for the pump relays. The first cycle will use these setpoints. Subsequent cycles rotate the setpoints among the pumps.
P112 2 0.3m P112 3 0.2m
Parameter Index Value Description P113 0 1.3m By using index 0 all six relays are set at the same time, including any alarm relays. Use with caution.
Starting Pumps by Rate of Level Change [MR 200]
Use this function when multiple pumps will be controlled by rate of level change rather than setpoints. Pumping costs can be reduced because only the highest ON setpoint needs to be programmed. This results in a lower difference in head to the next wet well which, in turn, results in less energy being used to pump out the well.
Parameter Index Value Description P112 1 1.35 Starting pumps by rate allows all setpoints to be set higher to save money by pumping from the highest safe level of the wet well.
P112 2 1.35 P112 3 1.35 P113 1 0.5m P113 2 0.5m P113 3 0.5m P121 1 1 P121 2 1 P121 3 1 P132 G 20.0
The pumps will start on 20 second intervals until the rate set in P703 is met.

When the first ON setpoint is reached, the pumps will start, one by one, until the material level rate of change is set at the same value or greater than the value in:
Set delay between pump starts using P132 – Pump Start Delay.
Prerequisite: Set pump relays to a service ratio value (P111 = 54 or 55).
Parameter Index Value Description P122 1 1 These values will start Pump Two 50% of the time and Pumps One and Three each 25% of the time. P122 2 2 P122 3 1
When more than one pump is assigned a Pump Service Ratio value (in any time units) and a pump start is required (P112 Relay Setpoint ON), the pump with the fewest running hours (with respect to the assigned ratio values) starts.
Conversely, when a pump stop is required (113 Relay Setpoint OFF), the pump with the most running hours (as compared to the assigned ratio values) stops.

Prerequisite: the volume of the vessel must be known.
Parameter Index Value Description P001 G 7 Operation = pumped volume P002 G 1
P003 G 2 P004 G 102 P005 G 1 P006 G 1.8 P007 G 1.4 P050 G 1 Tank shape is Flat Bottom. P051 G 17.6 Max volume is 17.6m3 or 17,600 liters. P111 1 52 Sets relays 1, 2, and 3 as a pump group using ALTERNATE DUTY ASSIST CONTROL. P111 2 52 P111 3 52 P112 1 1.0 Sets the ON setpoints for the pump group. P112 2 1.2 P112 3 1.4 P113 0 0.2 Sets the OFF setpoints for the pump group.
These parameters are as shown above.
Pump Control
Setting Independent Failsafe Controls
Independent failsafe controls allow you to vary an individual relay from the global failsafe controls programmed in P070 to P072.
Example:
The global failsafe controls are set to hold and Relay Five is set to trigger an alarm bell.
Parameter Index Value Description P071 G HOLd Keep level at last known value. P129 5 dE De-energize Relay Five, and trigger alarm.
![Page 76 of the Siemens 200 user manual — Reducing Wall Cling [MR 200]](/cdn/reader/82/829c15aaebfbf17c854075ac1e0dbc01f9b82abbfd8c8576c5c8e04c8c601513/images/page-76.webp?w=400)
When you need to pump below the normal OFF setpoint, use P130 (Pump Run-On Interval) and P131 (Pump Run-On Duration) to control this event.
Example:
The pump connected to Relay Three is set to pump for an extra 60 seconds every 5 hours.
Parameter Index Value Description P130 3 5 Time in hours of run-on interval. P131 3 60 Run-on for 60 seconds.
The pump start delay ensures that all of the pumps do not start at once to avoid power surges. There are two parameters used here: P132–Pump Start Delay and P133–Pump Power Resumption Delay. The default is 10 seconds but you can increase this if your pumps take longer to spin up.
Example:
The delay between pumps is set to 20 seconds and the delay of the first pump is set to 30 seconds.
Parameter Index Value Description P132 G 20 Wait at least 20 seconds between pump starts. P133 G 30 Wait for 30 seconds when power is restored.
Use the Wall Cling parameter to randomly alter the ON and OFF setpoints over a range. This eliminates the ridge of material that builds up at the setpoint that can give false echoes.
This setting may increase the number of days between trips to clean the wet well.
Wall cling reduction is set by P136. The relay setpoints ON and OFF are randomly varied inside a range so the material level does not stop at the same point.
![Page 77 of the Siemens 200 user manual — Setting a Flush Valve [MR 200]](/cdn/reader/82/829c15aaebfbf17c854075ac1e0dbc01f9b82abbfd8c8576c5c8e04c8c601513/images/page-77.webp?w=400)
Example [MR 200]:
A range of 0.5 meters is used to vary the setpoint. The randomly-selected setpoints are always inside the ON and OFF setpoints.
![Siemens 200 — Grouping Pumps [MR 200] diagram (page 77)](/cdn/reader/82/829c15aaebfbf17c854075ac1e0dbc01f9b82abbfd8c8576c5c8e04c8c601513/images/fig-77-1.webp)
You can group pumps and use the same pumping algorithm separately on each group. If you specify different pumping algorithms then the pumps are already grouped by algorithm and you do not need to use this parameter.
Group pumps only when four pumps are using the same algorithm and you want to split them into two groups.
Example:
Pumps One and Two can operate as a group and Pumps Three and Four can operate as another group.
Parameter Index Value Description P137 1 1 Groups Pumps One and Two P137 2 1 P137 3 2 Groups Pumps Three and Four P137 4 2
A flush valve stirs up the sediment on the bottom of the well during pumping so that it doesn’t accumulate. These parameters will control any relays set with P111 = 64 (Flush Valve).
Most sets of parameters will work with only one or two changes; however, for these parameters to work, all of them must be set to a value.

Example:
The flush valve connects to Relay Four and the watched pump is on Relay One.
Parameter Index Value Description P170 G 1 Watch Relay One to count pump cycles. P171 G 3 Open the flush valve for 3 cycles. P172 G 10 Use the flush value every 10 cycles. P173 G 120 Open the flush valve for 120 seconds.
A relay can be controlled directly by a remote system through communications. No other control schemes can then be used with a relay configured this way. Communications can be used to force status of some control relays, such as pumps.
Settings:
Parameter Index Value Description P111 5 65 Sets Relay Five to communications control.
You can find out how much an individual pump has been used by viewing the pump records parameters.
Information Available Parameter Access Current RUN time P309 Total pump hours P310 Total pump starts P311 Total pump RUN on occurrences P312 [MR 200 ONLY]

This feature is only available on the MultiRanger 200.
Screens or rakes are mounted on the inflow channel of the wastewater treatment plant to prevent debris from clogging the equipment.
When material builds up on the screen, a level differential is created, and the water level is higher in front of the screen than behind it. When this differential reaches the programmed setpoint, the MultiRanger activates a relay to operate mechanical rakes that clean the screen and ensure a steady flow.
P007 (3) Max differential between Point 1 and Point 2 reading also sets 100% scale for bargraph and mA output.



Prerequisite: Substitute the details of your application in place of the sample values provided. If you are bench testing the unit, set your test values to be the same as the sample values.
Parameter Index Value Description P001 G 4 Operation = Differential P002 G 1 Material = liquid P003 1,2 2 Max. Process Speed = medium P004 1,2 102 Transducer = XPS-10 P005 G 1 Units = meters P006 1 2 1.8 2.2 Empty Empty
= 1.8m = 2.2m P007 1 2 3
1.4 1.4 1.4
Span Span Max Differential
= 1.4m = 1.4m = 1.4m
Parameter Index Value Description P110 1 3 Starts the rake when the difference between the two levels rises above 0.4m and stop the rake when the difference falls below 0.1m.
P111 1 50 P112 1 0.4 P113 1 0.1
P110 2 1 Description P110 2 1 Sets Relay Two as a high level alarm for Transducer One with an ON setpoint of 1.3m and an OFF setpoint of 1.2m.
P111 2 1 P112 2 1.3 P113 2 1.2 P110 3 2 Sets Relay Three as a low level alarm for Transducer Two with an ON setpoint of 0.2m and an OFF setpoint of 0.4m.
MR 200
P111 3 1 P112 3 0.2 P113 3 0.4 P110 4 3 Sets Relay Four as a rake failure alarm as it uses the differential level point (3) with an ON setpoint of 1.0m and an OFF setpoint of 0.9m.
P111 4 1 P112 4 1.0 P113 4 0.9

This feature is only available on the MultiRanger 200.
External totalizers are simple counters which count the number of relay clicks produced by the MultiRanger. This is generally used to keep track of OCM or pumped volume totals. Note that both of these values are also stored in the MultiRanger and are available through communications.
Flow samplers are devices which take a sample of liquid when triggered by a relay click. These samples are used to monitor water quality over time. Flow samplers can be driven by OCM volume or by relay click volume settings depending on the application requirements.
Pumped volume is calculated at the end of the pump cycle. Totalized volume given through a relay set up for totalizer (P111[r]=40) will be given in bursts at this time.
Both the open and closed times for the relay contact are provided by P645 and are preset to 0.2 seconds. Partial units are added to the next pump cycle.
Example:
Shows a relay set up to make one contact for every cubic metre (m3) of liquid.
4m3 2m3 2m3 3m3 1m3
To set the totalizer to provide relay contact to an external counter, use the following:
1 Contact per 10P640 units P640 is preset to 0 so the default number of contacts for a pumped volume cycle is equivalent to the number of volume units.
Pump Cycle

Operation Units Source Parameter OCM (P001=6) P604–Maximum Flow, or P608–Flowrate Units Pumped Volume (P001=7) P051–Max Volume
Based on Volume and Time
To trigger a flow sampler relay based on flow, use P111[r]=41 and set the other parameters:
1 Contact per P641 x 10P642 units
Operation Units Source Parameter OCM (P001=6) P604 – Maximum Flow, or P608 – Flowrate Units
By using a mantissa (P641) and an exponent (P642), the relay contacts can be based on a volume other than a multiple of ten.
During the periods of low flow, the sampler may be idle for lengths of time. Program P115 to a time interval in hours to drive the sampler. The sampler will operate based on the volume of flow or the time interval, whichever comes first.
Flow Samplers – MR 200

An OCM installation is defined one of three ways, based on the Primary Measuring Device (PMD):
For some common weir and flume types. PMD dimensions (P602) are entered directly.
For most other weir and flume types. PMD exponents provided by the manufacturer are entered. Flow is calculated using the exponent (P601) and the maximum values (P603 and P604).
For all other PMDs, the head-to-flow curve can be plotted based on known breakpoints, usually supplied by the PMD manufacturer.
These Quick Start parameters are required for all installations.
Parameter Index Value Description P001 G 6 Operation = OCM P002 G 1 Material = liquid P003 G 2 Max. Process Speed = medium P004 G 102 Transducer = XPS-10 P005 G 1 Units = meters P006 G 1.8 Empty = 1.8m P007 G 1.0 Span = 1.4m P801 G 0.8 Range Extension to avoid LOE

Many PMDs start flowing higher than the traditional empty distance of the application. You can account for the flow in one of two ways:
P007
P603
P605
and above the bottom of the channel. It should be used if the surface monitored can fall past the Empty (P006) level in normal operation without reporting an LOE. The value is added to Empty (P006) and can be greater than the range of the transducer.
P007
P603
P801
The examples on the following pages show both methods.
m

To display the totalized volume on the LCD use the following parameters:
Parameter Index Value Description P737 G 2 Show the eight digit totalizer in the primary display

4 to 5 x hmax (hmax = P007)

Parameter Index Value P600 G 7–ISO 1438/1 V Notch Weir P602 1 Notch angle (view only) 2 Discharge coefficient (Ce) P603 G Maximum Head (preset to P007) P801 G Range Extension P608 G Flowrate Units
Front View

MR 200

Plan View 4 to 5 x hmax (hmax = P007)
Front View Side View


Parameter Index Value P600 G 6–ISO 4359 Rectangular Flume P602 1 Approach width (B) 2 Throat width (b) 3 Hump Height (p) 4 Throat length (L) (view only) 5 Velocity coefficient (Cv) (view only) 6 Discharge coefficient (Cd) (view only) 7 Cross sectional area P605 G Zero Head P608 G Flowrate Units
m · Transducer · P605 (p)

Transducer D/2, point of measurement
P007
P603
P605
Parameter Index Value P600 G 2–Palmer Bowlus Flume P602 1 Flume width (D) P603 G Maximum Head (preset = P007) P604 G Maximum Flow P605 G Zero Head P606 G Time Units

MR 200
P006 P007
D
Front View Side View
Parameter Index Value P600 G 3–H Flume P602 1 Flume height (D) P603 G Maximum Head (preset = P007) P604 G Maximum Flow P606 G Time Units
Point of Measurement cm inches 0.5 5 1¾ 0.75 7 2¾ 1.0 9 3¾ 1.5 14 5½ 2.0 18 7¼ 2.5 23 9 3.0 28 10¾ 4.5 41 16¼

For Primary Measuring Devices (PMDs) that measure flow by an exponential equation, use these parameters. Ensure that you use the correct exponent for your PMD; the values below are samples only.


Parameter Index Value P600 G 1 - Exponential Function P601 G Weir Type Value1
2.50 1.50 1.50 1.00 P603 G Maximum Head P604 G Maximum Flow P606 G Time Units P801 G Range Extension
V-notch Suppressed rectangular Cipolletti or trapezoidal Sutro or proportional

MR 200
compound Poebing approximate exponential
Flows through these weirs can be measured using the Universal Flow Calculation P600 = 4 or 5. See Universal Calculation Support on page 83.
P006
Application Information
Parameter Index Value P600 G 1–Parshall Flume P601 G 1.22–1.607 (consult your flume documentation) P603 G Maximum Head P604 G Maximum Flow (Q) P606 G Time Units

Parameter Index Value P600 G 1–Leopold Lagco Flume P601 G 1.55 P603 G Maximum Head (preset P007) P604 G Maximum Flow P605 G Zero Head P606 G Time Units

Flume Size (pipe diameter in inches)
Point of Measurement cm inches 4-12 2.5 1 15 3.2 1¼ 18 4.4 1¾ 21 5.1 2 24 6.4 2½ 30 7.6 3 42 8.9 3½ 48 10.2 4 54 11.4 4½ 60 12.7 5 66 14.0 5½ 72 15.2 6
Cut Throat Flume
Application Information
Parameter Index Value P600 G 1–Cut Throat Flume P601 G 1.55 P603 G Maximum Head (preset P007) P604 G Maximum Flow P606 G Time Units
m · Plan View

When the primary measuring device (PMD) doesn’t fit one of the standard types, it can be programmed using a universal characterization. When Universal is selected as the PMD type (P600), then both P610 and P611 must be entered to define the flow.
Two curve types are supported:
Both are shown in the following chart.
Characterization is achieved by entering the head (P610) and corresponding flow (P611), either from empirical measurement or from the manufacturer's specification. Increasing the number of defined breakpoints will increase the accuracy of the flow measurement.
Breakpoints should be concentrated in areas exhibiting the higher degrees of non linear flow. A maximum of 32 breakpoints can be defined. The curve’s end point is always specified by the parameters Maximum Head (P603) and Maximum Flow (P604) for a maximum total of 33 breakpoints.
Max. Flow
Max. Head
Use as many breakpoints as required by the complexity of your PMD. See Volume on page 51 for more information and parameters P610 and P611 for characterization.

These example flumes would both require a universal calculation.
Trapezoidal
Dual Range (nested) Parshall
Example Weirs
These weirs could require universal calculation.
compound Poebing approximate exponential
m · contracted rectangular

After programming the unit, you must test the device to ensure that it performs to your specifications. This test can be run in simulation mode or by varying the level in the application. The latter is preferred as it more accurately represents running conditions. However, if it is not possible to do a physical test, a simulation will ensure that control programming is correct.
In simulation mode, the LCD display reacts to the simulated level changes. Alarm relays will also react to the simulation, but any pump or control relays will not react.
To allow pump or control relays to operate on the simulated level, set P000 to –1.
Access the appropriate parameter: (press PROGRAM and then key in the parameter number). Press TRANSDUCER five times to overcome Echo Lock (P711) if applicable: the associated Reading is displayed in the Parameter Value field, and any alarm relays are set accordingly.
1. Go to the parameter to be simulated (P920, P921, P922, P923, P924, P925, or P926), and key in a material level in Units (Units defined in P005), or % of Span (% of Span defined in P007). 2. Press ENTER to display the calculated Reading. [Regardless of setting for P001, the parameter value being simulated will display as the primary reading. The auxiliary reading will always display the level value (P921). See “Display” on page 22 for positions of the primary and auxiliary readings on the LCD.] 3. Verify the calculated Reading. (At this point, the simulation is in Stop state. The
effect of the ARROW key is based on this state. See chart on page 86.)

The effect of the ARROW key is determined by the state (rate of rise or fall) immediately before the key is pressed.
Action State (prior to pressing key) Effect
Stop Rise at 1% of Span / second
Rise at 1% of Span / second Rise at 4% of Span / second (max.)
Rise at 4% of Span / second (max.) No effect
Fall at 1% of Span / second Stop
Fall at 4% of Span / second Fall at 1% of Span / second
Stop Fall at 1% of Span / second
Rise at 1% of Span / second Stop
Rise at 4% of Span / second (max.) Rise at 1% of Span / second
Fall at 1% of Span / second Fall at 4% of Span / second (max.)
Fall at 4% of Span / second (max.) No effect
When the level rises to 100% or falls to 0%, it reverses direction at the same rate.
Press · Press

After the unit is installed, test to verify the wiring.
Use P119 to force a state change and verify that the results are as expected (pump starts, alarm sounds, etc.).
Discrete Inputs
Use P270 to force the input value and verify that the results are as expected.
1. Go to P270 [DI] where DI = the discrete input to be tested 2. Set P270 to 0 (forced OFF) 3. Go to P275 [DI] to verify that the value is forced 4. Check the state of outputs to ensure that they respond as expected 5. Go to P270 [DI] 6. Set P270 to 1 (forced ON) 7. Go to P275 [DI] to verify that the value is forced 8. Check the state of outputs to ensure that they respond as expected
For further information see Discrete Inputs section on page 48.
Use P254 to test the mA input value against a true level. Use a trusted external mA source to generate the signal required for testing, and verify the incoming signal with P260. Check that the system responds as expected when the mA level is changed.
Use an external device to test the mA output against the measured level. Check that the mA value changes to reflect the changes in the measured level.
If you are testing the application by varying the material level (the preferred test method) make sure that none of the control devices is connected (or at least that no power is available to them).
If you are testing the application in simulation mode (and P000 is not –1), then control relays are not energized and the control devices can remain connected.
While the level is being cycled, check the results of the discrete inputs either by closing the circuit externally (preferred) or by using P270 Discrete Input Function to force the input ON or OFF. Try all possible combinations to thoroughly test the setup. For each combination, run a complete cycle to verify that the relays operate as expected.

Monitor system performance carefully, under all anticipated operating conditions.
parameters for the new functionality.
Troubleshooting, on page 231.
If you cannot observe all possible operating conditions during the System Performance Evaluation, use the level simulation (see page 85) to verify programming.
When a simulation is run, alarm relays will react to the simulated level changes, but control relays will not react. You can set P000 to value –1 to trigger the control relays based on the simulated level.
Retest the system every time you adjust any control parameters.

MultiRanger Communication Systems
The MultiRanger is an integrated level controller capable of communicating process information to a Supervisory Control and Data Acquisition (SCADA) system, via a serial device such as a radio modem, leased line, or dial-up modem.

Connection via radio modem, dial-up modem, or leased line modem
The standard MultiRanger supports the following two communication protocols:
Modbus is an industry standard protocol used by SCADA and HMI systems. The MultiRanger uses Modbus to communicate via the RS-485 port. For a description of the Modbus protocol, contact your local Schneider representative.
Dolphin is a proprietary Siemens Milltronics protocol designed to be used with Dolphin Plus. For more information on Dolphin Plus or to obtain a copy of the software, please go to www.siemens.com/processautomation to contact your Siemens Milltronics representative.
The standard MultiRanger unit may also be enhanced with Siemens Milltronics SmartLinx® communication modules that provide an interface to popular industrial communication systems.
This manual only describes the built-in communications. For more information on SmartLinx, please consult the appropriate SmartLinx manual.
MultiRanger

The MultiRanger is capable of communicating with most SCADA systems, PLCs, and PCs. The supported protocols are:
The MultiRanger comes with two communication ports on the base unit.
Port Connection Location Interface 1 RJ-11 connector inside enclosure on main board RS-232 2 terminal block terminal block RS-485
The RJ-11 jack connects to a laptop computer for the following:
The RS-485 port on the terminal blocks connects into industrial communications wiring and has the following advantages:
To communicate with equipment requiring RS-485 capability, Siemens Milltronics offers the RS-485 External Modem Kit. For more information, go to www.siemens.com/processautomation
The Modbus protocol is supported in the base unit and can be configured using the Communications parameters P770 to P782.
To set up communications with a Modbus RTU master device on port 2 using RS-485, set the following parameters:

Parameter Index Value Description P770 2 3 Modbus RTU slave P771 2 1 Network address, only used for RS-485 P772 2 9.6 Data rate of 9600 baud P773 2 0 No parity, common setting P774 2 8 8 data bits, common setting P775 2 1 1 stop bit, common setting P778 2 0 No modem connected P782 2 0 Index parameter values globally
Other protocols are available through optional SmartLinx communications modules. Details on how to install and program these modules are contained in the SmartLinx documentation.
Dolphin Plus software makes it easy to record and compare parameter sets for all the MultiRangers in your company. Dolphin Plus uses a proprietary protocol called Dolphin to communicate with Siemens Milltronics instruments. This protocol is set when P770 = 1.
By default the settings for port 1 (RJ-11 connection) and Dolphin Plus match. These settings are:
Parameter Index Value Description P770 1 1 Dolphin P772 1 115.2 Data rate of 115.2 Kilo baud P773 1 0 No parity, common setting P774 1 8 8 data bits, common setting P775 1 1 1 stop bit, common setting

Wiring Guidelines

Port Wall Mount 1 RS-232 port (RJ-11 modular telephone jack) is on the motherboard and is generally used with a laptop computer or modem. 2 Connections for the RS-485 port are on the terminal block.
The RJ-11 jack and the RS-485 port are inside the enclosure of the unit.

RJ-11 Jack · RS-485 Port

To connect the unit to a PC using an RS-232 jack, use the cable as shown:

Modular cable with RJ-11 plug

Modular cable with RJ-11 plug
MultiRanger Terminal Block DB 9 DB 9

Computer DB-9 (male) · RS-232 Modem

(Shield to be grounded at one end)
To connect the unit to a modem using an RS-485 port:
COM MultiRanger Terminal Block RS-485 Modem
B
Configuring Communication Ports (Parameters)
The 11 parameters listed are indexed to the two communication ports, unless otherwise noted: An asterisk (*) identifies the preset value.
Port Description 1 RS-232 port (RJ-11 modular telephone) 2 The RS-485 port is on the terminal blocks
The communications protocol used between the MultiRanger and other devices.
Primary Index Communications Port
0 Communications port disabled 1 * Siemens Milltronics Dolphin protocol (preset for port 1) 2 Modbus ASCII slave serial protocol 3 * Modbus RTU slave serial protocol (preset for port 2)
The MultiRanger supports the Siemens Milltronics Dolphin format (www.siemens.com/ processautomation), and the internationally recognized Modbus standard in both ASCII and RTU formats. Other protocols are available with optional SmartLinx cards.
The unique identifier of the MultiRanger on the network.
Primary Index Communications Port Values Range: 0 to 9999 1 * Preset
MultiRanger Terminal Block · Values

For devices connected with the Siemens Milltronics protocol, this parameter is ignored. For devices connected with a serial Modbus slave protocol, this parameter is a number from 1-247. The network administrator must ensure that all devices on the network have unique addresses. Do not use the value 0 for Modbus communications as this is the broadcast address and is inappropriate for a slave device.
The communication rate with the master device.
Primary Index Communications Port
4.8 4800 baud 9.6 9600 baud 19.2 * 19,200 baud (preset for port 2) 115.2 * 115,200 baud (preset for port 1)
This specifies the rate of communication in Kbaud. Any value may be entered, but the only values supported are those shown above. The baud rate should reflect the speed of the connected hardware and protocol used.
The serial port parity.
Primary Index Communications Port
0 * No Parity 1 Odd Parity 2 Even Parity
Ensure that the communications parameters are identical between the MultiRanger and all connected devices, as many modems default to N-8-1.
The number of data bits per character.
Primary Index Communications Port
Range: 5 to 8 8 * Modbus RTU 7 or 8 Modbus ASCII 7 or 8 Dolphin Plus
P775 Stop Bits
The number of bits between the data bits.
Primary Index Communications Port Values Range: 1 or 2 1 * Preset
Values · Values · Values

Sets the MultiRanger to use an external modem.
Primary Index Communications Port
0 * No modem connected 1 Answer only
Sets the time that the unit will keep the modem connected with no activity.
Primary Index Communications Port Values Range: 0-9999 seconds
0 * No timeout
To use this parameter, ensure that P778 (Modem Available) =1. Ensure that the value is low enough to avoid unnecessary delays when an unexpected disconnect occurs but long enough to avoid timeout while you are still legitimately connected. This parameter value is ignored by the Modbus Master Drivers as they automatically disconnect when done communicating.
If the line is idle and the P779 Modem Inactivity Timeout expires, then the modem is directed to hang up the line. Ensure that P779 is set longer than the standard polling time of the connected master device. Set P779 to 0 to disable the inactivity timer.
Determines where index information is stored for the parameter access area for the Modbus register map.
Primary Index Global Values 0 * Global 1 Parameter-Specific
For more on Parameter Index Location, see Parameter Access on page 102.
Values

Features affecting volume, mA input, and Average or Difference readings apply to the MultiRanger 200 only. They are clearly marked.
The memory map of the MultiRanger occupies the Modbus holding registers (R40,001 and up). This map is used when the protocol is Modbus RTU slave or Modbus ASCII slave.
Legend Type The type of data held in the group of registers. Start The first register to hold the referenced data. Data Type The possible values of the data in the register. See Data Types on page 107 for more information. Description The type of data held in the individual registers. #R The number of registers used for the referenced data. Read/Write Indicates whether the register is readable, writeable or both.
Data Type Read/ Write Word Order 40,062 0/1 R/W
Map ID Register Map Type 40,063 1 0/1 = P782 R/W ID Siemens Milltronics Product Code
40,064 1 4 = Model 200 6 = Model 100 R
Single Parameter Access (SPA) R40,090 7 see Appendix A on page 114
41,010 2 -20,000 to 20,000 R
Volume (2)3[MR 200]
41,020 2 -20,000 to 20,000 R
2. Varies according to model. MR 100: Available as reading 1 and reading 2 when in either Single or Dual Point Mode. In Dual Point Mode, reading 1 and reading 2 are always available. MR 200: Available as reading 1, reading 2, and Average or Difference when in either Single or Dual Point Mode. In Single Point Mode, point 2 and 3 are only available if P001 = Average or Difference. In Dual Point Mode, reading 1 and reading 2 are always available. Point 3 is only available if P001[3]= Average or Difference.
Point Data · Reading (3)2

Data Type Read/ Write Point Data
Temperature (2) 41,030 2 –50 to 150 R Totalizer for points 1 and 2 [MR 200]
Discrete Inputs (2) 41,070 1 Bit Mapped R Relay Outputs (3 or 6) 41,080 1 Bit Mapped R/W
mA Input (1) [MR 200]
41,090 1 0000 to 20,000 R
mA Output (2) 41,110 2 0000 to 20,000 R/W
Pump on Setpoint (3 or 6)
41,420 6 0000 to 10,000 R/W
Pump off Setpoint (3 or 6)
41,430 6 0000 to 10,000 R/W
Pumped Volume (2) [MR 200]
Pump Hours (3 or 6) 41,450 12 UINT32 R Pump Starts (3 or 6) 41,470 6 0000 to 10,000 R
Parameter Access 43,998 to 46999 R/W
The MultiRanger was designed to make it easy for master devices to get useful information via Modbus. This chart gives an overview of the different sections. A more detailed explanation of each section follows below.
This determines the format of unsigned, double-register integers (UINT32).
See Unsigned Double Precision Integer (UINT32) on page 107 for more information.
This value identifies the register map used by the MultiRanger. See P782 Parameter Index Location on page 96.
See also Parameter Access (R43,998 – R46,999) on page 102 for details.
Pump Control

This value identifies the Siemens device type:
Device Type Value MultiRanger 100 6 MultiRanger 200 4
Measurement point data contain the current instrument readings. These are the values shown for the reading measurement for each measurement point. The reading is based on the setting for P001 (operation). P001 can be set to level, distance, OCM flow, or volume. See Parameter Reference section on page 117 for details.
The measurement registers are 41,010 to 41,012. The MultiRanger uses 41,010 when configured with a single transducer and 41,010 to 41,012 when configured with two transducers (P111=4 or 5 only). Two transducers can create three readings because they can generate an average or differential reading (R41,012) as well as the two level readings (R41,010 and R41,011).
Data Registers Parameter Reading 41,010 to 41,012 P920 Volume [MR 200] 41,020, 41,021 P924 Temperature 41,030 and 41,031 P912
The reading is expressed as a percentage of full scale, multiplied by 100:
Reading Value 0 0.00% 5000 50.00% 7564 75.64% 20,000 200.00%
The totalizers are stored as 32 bit integers using two registers. The totalizers can be read with R41,040 and R41,041 as totalizer for Point 1, and R41,042 and R41,043 as totalizer for Point 2. The totalizer values can be reset to any value by writing that value to the registers. The values can be cleared by writing zero (0) to the registers.
Input/Output (R41,070 – R41,143)
The MultiRanger has discrete inputs, mA inputs, mA outputs and relay outputs. See below for details for each I/O type.

This table shows the current status of the discrete inputs. Only register 41,070 is used.
Discrete Input Data Address 1 41,070, bit 1 2 41,070, bit 2
This table shows the current status of the relays. A reading of 0 means that the relay function is not asserted and a 1 means that it is asserted. For example, a 1 for a pump relay means that the pump is running.
Relay Data Address 1 41,080, bit 1 2 41,080, bit 2 3 41,080, bit 3 4 41,080, bit 4 5 41,080, bit 5 6 41,080, bit 6
Values are written to control a relay only if the Relay Control Function (P111) is set to communications (65). See Relay Function Codes (P111 Only) on page 110.
The mA input is scaled from 0 to 2,000 (0 to 20 mA multiplied by 100). P254 displays the value of the input. This parameter is indexed by the input number.
The mA output is scaled from 0 to 2,000 (0 to 20 mA multiplied by 100). This is displayed in P911.
Only relays set for pump control (P111 = 50 to 52) are available. These registers have no effect on relays programmed for other uses.
The ON setpoint level (P112) for the referenced pump relay.
The setpoint is scaled from 0 to 10,000 (0 to 100% of span multiplied by 100). So 54.02% is shown in the register as 5402.

The OFF setpoint level (P113) for the referenced pump relay.
The setpoint is scaled from 0 to 10,000 (0 to 100% of span multiplied by 100). So 54.02% is shown in the register as 5402.
The pumped volume registers hold the current total for all of the pumps associated with a level point. These registers are available only if operation is set to pumped volume (P001 = 7).
These volumes can become very large. Therefore, two registers are used to hold the value. See Unsigned Double Precision Integer (UINT32) on page 107 for more information.
The value in the registers is given as an integer value but must be interpreted as having the number of decimals set in P633 (LCD Totalized Decimal Position): this number can be 0 to 3. Ensure that your software accounts for these decimal places before you report the pumped volume totals.
The number of running hours for the referenced pump relay. The hours are given to three decimal places, so the integer must be divided by 1000 to get the correct value. For example 12,340 represents 12.34 hours.
This value comes from parameter P310. See page 157 of the Parameter Reference section for details.
The number of pump starts for the referenced pump relay.
This value comes from parameter P311. See page 157 of the Parameter Reference section for details.

Parameter values are given as integers in the range of registers from R44,000 to R44,999. The last three numbers of the register correspond to the parameter number.
Parameter Register # Format Register # Parameter # 44,000 46,000 P000 44,001 46,001 P001 44,002 46002 P002 … … … 44,999 46,999 P999
Usually, the parameters are all read / write.
Each parameter register has a corresponding format register that holds the format information required to interpret the value. See Format Words (R46,000 to R46,999) on page 105.
Many parameters are indexed. There are two possible indexes: a primary index and a secondary index. A secondary index is a sub-address of the primary index. Some indexed parameters affect multiple I/O devices.
The following is an example of a primary index:
P111 is the Relay Control Function. This parameter determines how a relay is controlled by the MultiRanger (used as an alarm, for pump control, etc.). Because there are up to six relays on the MultiRanger, P111 is indexed by six to allow each relay to be programmed independently.
A few parameters also have a secondary index. While a secondary index is important for setting up the MultiRanger, it is almost never needed through remote communications.
Each parameter communicates its value to only one register. You must know the index(es) for the parameter in order to interpret the information in the register correctly.
For example, to make use of the value returned in register R44,111 you must know which relay it is referring to. See Relay Function Codes (P111 Only) on page 110 for details on P111 values.
To determine the index values, the primary and secondary index must be read or write. The two possible methods of handling these index values are described in the following paragraphs: Global Index Method and Parameter Specific Index Method.

To read parameter values, follow the steps listed in either the Global or the Parameter Specific Index Method that follow. You must be able to program your HMI or SCADA system before completing these methods.
Global format method sets index values for all parameters simultaneously. Use this method to read multiple values set to the same index values.
This is a value between 0 and 40 which specifies the input or output indexed by the parameter.
This is a value between 0 and 40 that specifies the secondary index on the parameter. This value is usually 0.
primary and secondary indexes are already specified, these portions of the format word are ignored and only the last digit is significant.
See Format Registers on page 115 for details.
A value of 22,222 indicates that an error has occurred. Specify a different format type and try again.
The Parameter Specific index method sets the index values for each parameter independently. Use this method to read multiple parameters with different index values.

Send the integer value 01008 to register 46,921.
A value of 22,222 indicates that an error occurred. Specify a different format type and try again.
The method of writing parameters is similar to the method of reading them. Become familiar with Reading Parameters, page 103, before attempting to write any parameters.
To write parameter values to the MultiRanger, follow these steps:

Format words are unsigned integers that contain up to three values (described below). The number of values used in the format words depends on the Parameter Index Location (P782) that is used.
Parameter P782 Parameter Index Location, described on page 96, determines which of two methods is used to access the format words: Global Index Method or Parameter Specific Index Method.
Only the final digit of the format word determines the decimal offset (below).
All three decimal fields are used to determine the parameter value’s primary index, secondary index, and decimal offset.
Each format register is made up of three decimal fields:
The primary and secondary indices correspond to those that are used by the parameter.
decimal offset (0-9) · secondary index (00-99) · primary index (00-64)

The decimal offset indicates how the remote system must interpret the integer value that is stored in the parameter access register. The following table shows how different parameter values can be shown based on a register value (integer) of 1234.
Decimal Offset Example 0 0 1,234 1 –1 12,340 2 –2 123,400 3 –3 1,234,000 4 –4 12,340,000 5 –5 123,400,000 6 +1 123.4 7 +2 12.34 8 +3 1.234 9 Percent 12.34%
Examples of using the format word for both the index values and the decimal offset value are shown below:
00000 00 00 0 01003 01 00 3 right 02038 02 03 3 left 05159 05 15 percent
To write these values you can use a decimal offset as follows: format word = (primary index x 1000) + (secondary index x 10) + (decimal).

The MultiRanger parameters do not always use integers to hold values. For the convenience of the programmer, those values are converted to and from a 16-bit integer number. This section describes the conversion process. The sections that follow describe where those values are in the discrete I/O and block transfer addresses, and how to get the parameters you need.
Numeric parameter values are the most common. For example, parameter P920 (Reading) returns a number that represents the current reading (either level or volume, depending on the MultiRanger configuration).
Numeric values are requested or set in units or percent of span, and may be specified with a number of decimal places.
Numeric values must be in the range –20,000 to +20,000 to be valid. If a parameter is requested and its value is more than +20,000, the number 32,767 is returned; if it is less than –20,000, the number –32,768 is returned. If this overflow happens, decrease the number of decimal places.
If a parameter cannot be expressed in terms of percent of span, or has no meaningful value, the number 22,222 is returned. Try requesting the parameter in units, or refer to P005 in the Parameter Reference section on page 122.
Bits are packed into registers in groups of 16 bits (1 word). In this manual, the bits are numbered from 1 to 16, with bit 1 as the least significant bit (LSB) and bit 16 as the most significant bit (MSB).
16 15 14 13 12 11 10 09 08 07 06 05 04 03 02 01 MSB LSB
Large numbers are put into unsigned 32 bit integers. By default, they are set up so that the first word (register) is the most significant word (MSW) and the second word (register) is the least significant word (LSW).

For example, if R41,442 is read as a UINT32, the 32 bits would look like this:
Data Types
16 MSW 1 16 LSW 1
32 32-bit integer value (UINT32) 1
The two registers are read as a 32-bit integer.
The most significant word (MSW) and least significant word (LSW) can be reversed to accommodate some Modbus drivers. See Word Order (R40,062) on page 98 for details.
The position of the decimal place is dependent on the register. For more details see the description of the register.
Certain parameters are actually a pair of numbers separated by a colon, using this format: xx:yy.
xx = the average noise value in dB
yy = the peak noise in dB
The number which corresponds to xx:yy, either for reading or setting a parameter, is determined by the following formula:
For storing to the device:
value = (xx + 128) x 256 + (yy + 128)
For reading from the device:
xx = (value / 256) – 128 yy = (value % 256) – 128 where % is the modulus operator.
The modulus can be computed by following these steps:
value1 = value / 256 value2 = remainder of value1 value3 = value2 x 256 yy = value3 – 128
It may simplify Parameter to notice:
xx = (most significant byte of value) – 128 yy = (least significant byte of value) – 128

If a device parameter returns a text message, that message is converted to an integer and provided in the register. The numbers are shown in the following table:
Number Text Message as displayed on LCD 22222 Invalid value 30000 Off 30001 On 30002 ≡ 30003 (parameter does not exist) 30004 Err 30005 Err1 30006 Open 30007 Short 30008 Pass 30009 Fail 30010 Hold 30011 Lo 30012 Hi 30013 De 30014 En 30015 (parameter has not been set) -32768 Value is less than –20,000 32767 Value is greater than 20,000

Please note that the MultiRanger 200 offers more function codes.
If a device parameter returns a relay function code, that message is converted to a number and is then provided in the register. The numbers are shown in the following table:
Control Relay Function Code Number P111 General OFF, relay not used 0 0 Undesignated Level Alarm 1 1 Low-Low Level Alarm 2 1 – LL Low Level Alarm 3 1 – L High Level Alarm 4 1 – H High-High Level Alarm 5 1 – HH Loss of Echo (LOE) Alarm 20 6 Transducer Cable Fault Alarm 16 7 Pump Fixed Duty Assist 25 50 Alternate Duty Assist 30 52 Control Communication 66 65
Control Relay Function Code Number P111 General OFF, relay not used 0 0 Undesignated Level Alarm 1 1 Low-Low Level Alarm 2 1 – LL Low Level Alarm 3 1 – L High Level Alarm 4 1 – H High-High Level Alarm 5 1 – HH In Bounds Alarm 6 2 Out of Bounds Alarm 9 3 Rate of Level Change Alarm 12 4 Temperature Alarm 15 5 Loss of Echo (LOE) Alarm 20 6 Transducer Cable Fault Alarm 16 7 Pump Totalizer 22 40 Flow Sampler 23 41 Fixed Duty Assist 25 50 Fixed Duty Backup 26 51 Alternate Duty Assist 30 52 Pump (con’t)
Alternate Duty Backup 31 53 Service Ratio Duty Assist 35 54 Service Ratio Duty Backup 36 55 First In First Out (FIFO) 40 56 Control Flush Valve 65 64 Communication 66 65
See P111 on page 134 of theParameter Reference section.

Modbus Responses
When polled by a Modbus Master, a slave device will do one of the following:
for more details). This is the normal response.
MultiRanger uses the following exception codes:
Code Name Meaning 01 Illegal Function The function code received in the query is not an allowable action for the slave. 02 Illegal Data Address The data address received in the query is not an allowable address for the slave. 03 Illegal Data Value A value contained in the query data field is not an allowable value for the slave.
Errors can be traced to two general sources:
In the first case, the MultiRanger does not respond and the master waits for a response time out error, which causes the master to re-send the message.
In the second case, the response depends on what the host tries to do. In general, MultiRanger will not give an error to the host request. Various actions and the expected outcome are as follows:


Generally
computer used to communicate with the unit.
driver will solve the problem. An easy stand-alone driver called ModScan32 is available from Win-Tech at www.win-tech.com This driver is helpful for testing communications.
remains unchanged.

This Appendix is intended to provide someone with advanced communications knowledge the ability to access any parameter value in any available format.
Built into MultiRanger is an advanced handshaking area that can be used to read and write single registers to the MultiRanger. This section performs a similar function to the Parameter access section. The differences are:
Parameter Read and Write (40,090 – 40,097) is a series of eight registers used for reading and writing parameter values to and from the MultiRanger. The first three registers are always unsigned integers representing parameters and index values. The second five registers are the format and value(s) of the parameter.
All parameters normally accessed through the hand-held programmer are available through these registers.
Address Description 40,090 Parameter (integer) 40,091 Primary Index (integer) 40,092 Secondary Index (integer) 40,093 Format word (bit mapped) 40,094 Read value, word 1 40,095 Read value, word 2 40,096 Write value, word 1 40,097 Write value, word 2
Reading Parameters
To read parameters through Modbus do the following steps:
confirm that the operation is complete.

To set parameters through Modbus do the following steps:
interpret the value correctly.
Bits Values Description 1-8 0-2 Error Code 9-11 0-7 3-bit number representing decimal offset 12 0/1 direction of offset (0 = right, 1 = left) 13 0/1 Numeric format: Fixed (0) or Float (1) 14 0/1 Read or Write of data, Read (0), Write (1) 15 0/1 Word order: Most Significant Word first (0), Least Significant Word first (1) 16 Reserved
For example, to format the level reading so that it is shown in percent with two decimal places shifted left, the format bits would look like this:
Bit Numbers 16 15 14 13 12 11 10 09 08 07 06 05 04 03 02 01
Bit Values 0 1 0
offset direction to right
decimal offset of 2
The value sent to the MultiRanger is 0001001000000000 binary or 512 decimal. The value 512 is sent as an integer to register 40,093 to format the output words 40,094 and 40,095 accordingly.
If the numeric data type is set for integer and the value contains decimal places, they are ignored. In this situation, use the decimal offset to ensure that you have an integer value and then write your code to recognize and handle the decimal offset.

The error codes returned in the format area are 8-bit integers found in the lowest eight bits of the format word. This allows for 256 potential error codes.
Currently the MultiRanger has two error codes available.
Values Description 0 No error 1 Data not available as percent (available as units) 2-255 Reserved

The MultiRanger is configured through its parameters, and the application determines the parameter values which are entered into the unit.
Please check your value entries carefully before operating the MultiRanger to ensure optimum performance.
All MultiRanger 100 and MultiRanger 200 parameters are listed in this section. MultiRanger 200 parameters are identified separately. The parameter title includes this marking [MR 200], indicating that parameter applies to the MultiRanger 200 only.
As a general rule, parameters affecting volume apply only to the MultiRanger 200.
Please note the following:
The icon appears under the index field. 2. Enter the secondary index, and then enter the values to set the secondary index.


Secures the MultiRanger from changes via the handheld programmer.
1954 * OFF (programming permitted)
-1 Simulation Controls (relays energize based on simu- lated level)
other lock activated (programming secured)
Related • Simulation on page 85
Access this parameter directly (type 000) and enter any value (except 1954) to secure programming lock. To unlock the MultiRanger, access this parameter and enter 1954.
P001 Operation
Sets the type of measurement required for the application.
1 Level – how full the vessel is (MR 200: volume – P050)
2 Space – how empty the vessel is (MR 200: ullage – P050)
3 * Distance – distance from transducer to material
4 DPD – dual point difference [MR 200]
5 DPA – dual point average [MR 200]
6 OCM – flow rate in an open channel [MR 200]
7 Pump Totalizer – total pumped volume [MR 200]
Alters • P600 Primary Measuring Device

Please note that this feature is only available on the MultiRanger 200.
For Dual Point Difference (DPD) or Dual Point Average (DPA), the unit requires either two transducers of the same type, or one transducer and one mA input. If two transducers are used, all transducer parameters become indexed, and a third level point is calculated.
The points must be globally set to either 4 or 5 (as required). Point 3 becomes the calculated value as shown above. See Rake Control example on page 69.
To set a dual point MultiRanger for DPA or DPD functions, Point 3 must be set to either 4 or 5 (as required). Points 1 and 2 cannot be set to 4 or 5, but these points are used to calculate the value in point 3.
This table shows the available functions:
P001 [1] 1, 2, 3, 6, 7
P001 [2] 1, 2, 3, 6, 7
P001 [3] 4,5
P002 Material
Specifies material type.
Values 1 * Liquid or horizontal solid surface
2 Solid or angled surface
Alters • P830 TVT Type
Global Transducer

Determines level change reaction.
1 Slow (0.1 m/min)
2 * Medium (1 m/min)
3 Fast (10 m/min)
Use a setting just fast enough to keep up with your process. Slower settings provide higher accuracy. Faster settings allow for more level fluctuations.
Specifies the Siemens Milltronics transducer connected to the unit.
Primary Index Single Point Model Dual Point Model Global Transducer
0 * No transducer attached (preset for Dual Point) 1 ST-25 2 ST-50 100 STH 101 XCT-8 102 * XPS-10 (preset for Single Point Model) 103 XCT-12 104 XPS-15 112 XRS-5 250 mA input [MR 200]
(chart con’t next page)

Related (chart cont’d from previous page)
Specifies measurement units used for dimensional values.
Changing this value automatically changes the units displayed for many parameters. Existing values are converted and do not have to be re-entered.
Enter distance in units (P005) from the face of the transducer to the process empty point.
Alters • P007 Span
Altered By • P005 Units
Setting this value also sets Span (P007) unless Span was already set to another value. For distance operation (P001=3), Span is preset to Empty.
![Page 133 of the Siemens 200 user manual — Volume (P050 to P055) [MR 200]](/cdn/reader/82/829c15aaebfbf17c854075ac1e0dbc01f9b82abbfd8c8576c5c8e04c8c601513/images/page-133.webp?w=400)
Sets the range levels to be measured.
Span is preset for a value close to the maximum available. Enter a value reflecting maximum application range.
Always prevent the monitored surface from approaching within 0.3 m (1 ft) of the transducer face as this is the minimum blanking for most Siemens Milltronics transducers (some require more blanking – see your transducer manual).
Many other parameters are set as a percentage of span (even if they are entered in units). The values of these other parameters may change if the span is altered after installation and the other parameters are measured using a level determined upward from the Empty level toward the transducer face.
All volumes are based on span so it should be set for the maximum volume point if volume calculations are needed.
Please note that the volume parameters apply only to the MultiRanger 200.
Use these parameters to enable the MultiRanger to show readings based on vessel or wet well volume (rather than level)..
Primary Index Level · Alters · Related
![Page 134 of the Siemens 200 user manual — P050 Tank Shape [MR 200]](/cdn/reader/82/829c15aaebfbf17c854075ac1e0dbc01f9b82abbfd8c8576c5c8e04c8c601513/images/page-134.webp?w=400)
Enter the Tank Shape value matching the monitored vessel or wet well.
When Operation is LEVEL (P001 = 1), liquid (material) volume is calculated.
Alternatively, when Operation is SPACE (P001 = 2), remaining vessel capacity is calculated. In RUN mode, readings are displayed in percent of maximum volume. To convert readings to volumetric units, see Maximum Volume (P051).
![Siemens 200 — P050 Tank Shape [MR 200] diagram (page 134)](/cdn/reader/82/829c15aaebfbf17c854075ac1e0dbc01f9b82abbfd8c8576c5c8e04c8c601513/images/fig-134-1.webp)
![Page 135 of the Siemens 200 user manual — P051 Maximum Volume [MR 200]](/cdn/reader/82/829c15aaebfbf17c854075ac1e0dbc01f9b82abbfd8c8576c5c8e04c8c601513/images/page-135.webp?w=400)
![Siemens 200 — Volume (P050 to P055) [MR 200] diagram (page 135)](/cdn/reader/82/829c15aaebfbf17c854075ac1e0dbc01f9b82abbfd8c8576c5c8e04c8c601513/images/fig-135-1.webp)
For Readings in volumetric units (rather than percent), enter the equivalent vessel volume for Span (P007).
![Siemens 200 — P051 Maximum Volume [MR 200] diagram (page 135)](/cdn/reader/82/829c15aaebfbf17c854075ac1e0dbc01f9b82abbfd8c8576c5c8e04c8c601513/images/fig-135-2.webp)
Any volume units can be chosen because volume is calculated from empty to maximum span and is scaled according to the Tank Shape (P050) value.
![Siemens 200 — P051 Maximum Volume [MR 200] diagram (page 135)](/cdn/reader/82/829c15aaebfbf17c854075ac1e0dbc01f9b82abbfd8c8576c5c8e04c8c601513/images/fig-135-3.webp)

Dimension A as used in P050 Tank Shape.
Related • P050 Tank Shape
Enter one of the following:
Dimension L as used in P050 Tank Shape.
Related • P050 Tank Shape
Enter the following:
[MR 200]
When the tank shape is too complex for any of the preconfigured shapes, you can specify the volume based on segments. See “Accessing a Secondary Index” on page 117 for more information.
Related • P055 Volume Breakpoints
Enter the following:

(Universal Volume Calculation) [MR 200]
Each segment defined by the level breakpoints (P055) requires a volume so that the MultiRanger can make the level-to-volume calculations.

V = (1/3)πr2h V = πr2h
For more on Volume Characterization, go to page 52.
These parameters are used to:

Defines the maximum number of decimal places used on the LCD.
3 3 digits after the decimal point (limited by device resolution)
Cone Cylinder

Alters • P607 Flowrate Decimal
Altered by • P051 Maximum volume
Related • P920 Reading Measurement
In RUN mode, the decimal position adjusts to prevent the number of digits from exceeding the display capabilities. To keep the decimal place from shifting, reduce the number of decimal places to that shown at 100%.
Example:
If 100% is 15m, use two decimal places for sample readings of 15.00 or 12.15.
Multiplies the current value by the specified amount to allow for scaling.
Related • P920 Reading Measurement
Adds the specified value to the level reading, usually to reference the reading to sea level or another datum level.
Related • P920 Reading Measurement
The operation of the device is not affected by the Offset Reading. This value is used for display purposes only. All control measurements are still referenced to Empty.
Primary Index Level · Primary Index Level

Use this feature to override the ultrasonic reading by a discrete input such as a contacting point device. The ultrasonic reading will be fixed at the programmed switch level until the discrete input is released.
The ultrasonic device makes decisions based on the override values.
Sets the discrete input to act as the source for a level reading override.
Values 0 * OFF: No override.
This value is substituted for the current reading when the selected discrete input is enabled and activated.

Please note the following:
Example:
Transducer One is configured for a level measurement. Digital Input 2 is connected to a Hi Level Backup switch located a level of 4.3m.
P064 1 2
P065 1 4.3
When the level rises to 4.3m, and the switch is activated, the reading is forced to 4.3m. The reading stays at 4.3m until the switch is de-activated.
Defines the time used to calm (debounce) the override condition input. Set in seconds.
Holds the current password for P000. Select by typing in 069. You cannot scroll to this parameter.
Related • P000 Lock
This parameter is write-only, and can only be selected by entering 069. To change the password, unlock the device by entering the current password into P000. Then enter the new password into P069. To lock the device, enter a password other than the correct one in P000. While the device is unlocked, the password is visible in P000.
Primary Index Transducer · Related · Primary Index Global

P070 Failsafe Timer
The time for invalid measurements to elapse before Failsafe State activates.
Preset 10.00 minutes
Altered by • P003 Maximum Process Speed
Related • P129 Relay Failsafe
When modifying the preset value, set it short enough to protect the process but long enough to avoid false alarms. Only use No Delay (0.0 Minutes) for testing.
The material level reported when a Failsafe State is initiated.
Value in units or % (-50% to 150% of span)
HI Level goes to maximum span
LO Level goes to 0 span (Empty)
HOLd * Level remains at last reading
Select the Failsafe Material Level based upon the relay operation required during failsafe operation.

To enter a specific Failsafe Material Level within -50 to 150% of Span (P007), in Units (P005).
The way in which relay programming reacts to the failsafe level depends on P129 Relay Failsafe (page 140). By default:
Sets the speed the MultiRanger advances to and returns from the Failsafe Material Level.
1 * Restricted Advances to/from Failsafe Material Level as set by P003, P700 and P701.
2 Immediate Failsafe Material Level assumed right away
3 Fast Back Failsafe Level Advance is restricted, return is immediate
The MultiRanger has three or six relays (or digital outputs) used to control devices and alarms. While the number of devices is limited by the relays, all control functions are accessible through software and each parameter is indexed to the three or six relays. See the Relay section on page 41.
The MultiRanger makes standard applications easier to program by providing an extensive list of presets.
Primary Index Level · Values · Related

Each relay can be configured independently to take advantage of the MultiRanger’s advanced features and flexibility. Start with a preset application and then change the required parameters to make the task more efficient.
Each relay is triggered by one or more setpoints. The setpoints can be based on absolute level (P112, P113) or rate of change (P702, P703). Each control function specifies which setpoints are required.
Six preset applications to configure or bench test the unit.
1 Wet Well 1
2 Wet Well 2
Related • P001 Operation
Select an application that is similar to yours and change the parameters required. If none suit, then refer to P111 Relay Control Function on page 134.

Sets the level source on which the indexed relay is assigned to a measurement point.
Please note that Difference and Average settings are MultiRanger 200 features.
1 * Point # 1 = Transducer One
2 Point # 2 = Transducer Two
Point # 3 = Difference (P001=4) or Average (P001=5) [MR 200]
Altered by:
Please note that these features are only available on the MultiRanger 200.
Points 2 and 3 are available only if Operation is set for difference or average (P001 = 4 or 5).
Point 2 is always available, and Point 3 is available only if Operation is set for difference or average (P001 = 4 or 5).
Sets the control algorithm used to trip the relay.
Please note that parameter P111 values vary between the MultiRanger 100 and the MultiRanger 200.
Values See chart below
Altered by • P100 Preset Applications
Use zero 0 (preset) to disable control of the indexed relay.
Primary Index Relay · Values · Primary Index Relay

Values For P111 Control Type # 1
Relay Control
OFF 0* Relay set off, no action (preset) Level 1 Based on level setpoints ON and OFF Loss of Echo (LOE) 6 When echo is lost Cable Fault 7 When the circuit to a transducer is opened
Pump Fixed Duty Assist 50 At fixed ON and OFF setpoints and allows multiple pumps to run Alternate Duty Assist 52 At rotating ON and OFF setpoints and allows multiple pumps to run
Communication 65 Based on input from external communications. See Communications section on page 89 for further refer- ence.
Values For P111 Control Type # 1 Relay Control
Off 0* Relay set off, no action (preset) Level 1 Based on level setpoints ON and OFF
In Bounds 2 When level enters the range between ON and OFF setpoints
Out of Bounds 3 When level exits the range between ON and OFF set- points Rate of Change 4 Based on rate setpoints ON and OFF Temperature 5 Based on temperature setpoints ON and OFF Loss of Echo (LOE) 6 When echo is lost Cable Fault 7 When the circuit to a transducer is opened
Flow Totalizer 40 Every 10y units (P641-P645) Flow Sampler 41 Every y x 102 units (P641-P645) or time duration (P115)
General · General

Fixed Duty Assist 50 At fixed ON and OFF setpoints and allows multiple pumps to run or for rake control
Fixed Duty Backup 51 At fixed ON and OFF setpoints and allows only one pump to run Alternate Duty Assist 52 At rotating ON and OFF setpoints and allows multiple pumps to run Alternate Duty Backup 53 At rotating ON and OFF setpoints and allows only one pump to run Service Ratio Duty Assist 54 On service ratio at ON and OFF setpoints and allows multiple pumps to run Service Ratio Duty Backup 55 On service ratio at ON and OFF setpoints and allows only one pump to run First In First Out (FIFO) 56 As Alternate Duty Assist, resets the relay from stag- gered OFF setpoints
Flush Valve 64 Used to control a pump flushing device based on Flush Systems (P170 to P173)
Communication 65 Based on input from external communications. See Communications section on page 89 for further refer- ence.
Sets the process point at which the relay changes from its NORMAL state.
Altered by • P007 Span
For most applications, the relay is tripped at this point. For IN-BOUNDS and OUT-OF- BOUNDS alarms, it is the high point in the specified range. This parameter is set according to Span (P007) even when another reading, such as volume, is shown on the LCD.
![Page 147 of the Siemens 200 user manual — P116 Dead Band [MR 200]](/cdn/reader/82/829c15aaebfbf17c854075ac1e0dbc01f9b82abbfd8c8576c5c8e04c8c601513/images/page-147.webp?w=400)
Sets the process point at which the relay returns to its NORMAL state.
Altered by • P007 Span
For most applications, the relay is reset at this point. For IN-BOUNDS and OUT-OF- BOUNDS alarms, it is the low point in the specified range. This parameter is set to Span (P007), even when another reading, such as volume, is shown on the LCD.
The length of time in hours between starts.
Altered by • P100 Preset Applications
Related • P111 Relay Control Function
The distance above and below the bound alarm setpoints.

For IN-BOUNDS and OUT-OF-BOUNDS Relay Functions (P111 = 2 and 3 respectively), a dead band prevents relay chatter due to material level fluctuations at both the upper and lower setpoints.
Enter the dead band in either percent of span or units of measure (P005). The dead band value is applied both above and below the upper and lower bound setpoints as shown in the fig- ure.
Actual ON or OFF setpoint
Setpoint set in P112 or P113
Actual OFF or ON setpoint
The logic applied to relays to determine the contact open or closed state.
Pump or Control
Contact
2 * Positive Normally Closed Normally Open
3 Negative Normally Open Normally Closed
Related • P111 Relay Control Function
The relay contact operation is NORMALLY CLOSED for alarms and NORMALLY OPEN for controls. See P111 Relay Control Function for more information.
When power is cut to the MultiRanger, its relays fail to the following states:
3,6 Open or Closed1
To use relays 3 or 6 as general alarm indicators, set P118 to 3 – negative logic and wire the alarm for normally open operation. When an alarm event occurs (see below) or when power is cut, the circuit closes and the alarm activates.

In software, all relays are programmed the same way, with ON setpoints indicating when to change the relay contact state (open or closed). This parameter allows the reversal of the operation so that relay contacts can be NORMALLY CLOSED or NORMALLY OPEN. P118 is preset to 2 which is positive logic.
When P118 = 3 (negative logic), the operation for the indexed relay is reversed from normal.
Forces the relay control logic into an ACTIVATED or DE-ACTIVATED state.
0 * OFF - Control from MultiRanger algorithms
1 Activate relay control
2 De-activate relay control
This parameter tests site wiring and control logic programming. Forcing the relay to an activated or de-activated state is similar to the MultiRanger detecting an event and responding to it. Helpful in testing new installations and diagnosing control problems.
Please note that these parameters apply to the MultiRanger 200 only.
These parameters provide alternate ways of starting the pumps in the pump group. See the Pump Control section on page 58 for descriptions of the pump control algorithms.
Sets the pump relays to accept control by rate of level change once the first ON setpoint is reached.
Values 0 * OFF (pump by level)
1 ON (pump by rate)
Use this function when multiple pumps are to be controlled by rate of level change rather than by setpoints.

The delay between pump starts is set by P132 Pump Start Delay.
This only applies to any relays set to pump control (P111 = 50 to 56).
Selects pump usage based on the RUN time ratio rather than last used.
Related • P111 Relay Control Function
This parameter only relates to relays with P111 = 54 or 55.
To make this parameter useful, assign it to all of the pump relays. The number assigned to each pump relay represents the ratio applied to decide the next pump to start or stop.
P129 Relay Failsafe
Sets how individual relays react to a failsafe condition, to allow for more flexible programming.
OFF * Response governed by P071 Failsafe Material Level
HOLd For LAST KNOWN relay state retention
dE To have the relay de-energize immediately on failsafe
En To have the relay energize immediately on failsafe
Altered by • P071 Failsafe Material Level
Use this for operations independent of the Failsafe Material Level (P071).
Primary Index Relay · Primary Index Relay · Values
![Page 151 of the Siemens 200 user manual — P130 Pump Run-On Interval [MR 200]](/cdn/reader/82/829c15aaebfbf17c854075ac1e0dbc01f9b82abbfd8c8576c5c8e04c8c601513/images/page-151.webp?w=400)
Relay Failsafe is only available for the following relay functions (P111) and is not used for any other relay control function.
50 to 56 – all pump controls dE
Please note that these parameters apply to the MultiRanger 200 only.
These parameters only affect relays set to pump operation (P111 = 50 to 56).
Sets the number of hours between pump run-on occurrences.
Related • Advanced Pump Control Modifiers (P130 to P136)
To clear sediment in a pump-down wet well, run the pump after the normal OFF setpoint is reached to force some solid material through. This parameter sets the time between such events. Only the last pump running can run-on.
Primary Index Global
![Page 152 of the Siemens 200 user manual — P133 Pump Power Resumption Delay [MR 200]](/cdn/reader/82/829c15aaebfbf17c854075ac1e0dbc01f9b82abbfd8c8576c5c8e04c8c601513/images/page-152.webp?w=400)
Sets the number of seconds that the pump runs on.
Related • Advanced Pump Control Modifiers (P130 to P136)
Your pump capacity determines the amount of material that can be removed. Choose a value long enough to clean out the vessel bottom, yet short enough not to run the pump dry. Also be sure that this value does not overlap with P130 (Interval). The timing should look something like this:
P131 P132 P130 (Pump Run-On)
Sets the minimum delay (in seconds) between pump starts.
Value is divided by 10 in simulation mode.
Use this feature to reduce a power surge from all pumps starting at the same time. This delay determines when the next pump is permitted to start.
Sets the minimum delay before the first pump restarts after power failure.
This reduces the surge from the first pump starting immediately on power resumption. When this delay expires, other pumps will start as per P132.
![Page 153 of the Siemens 200 user manual — P137 Pump Group [MR 200]](/cdn/reader/82/829c15aaebfbf17c854075ac1e0dbc01f9b82abbfd8c8576c5c8e04c8c601513/images/page-153.webp?w=400)
Varies the upper and lower setpoints to reduce material buildup on the walls.
![Siemens 200 — P136 Wall Cling Reduction [MR 200] diagram (page 153)](/cdn/reader/82/829c15aaebfbf17c854075ac1e0dbc01f9b82abbfd8c8576c5c8e04c8c601513/images/fig-153-1.webp)
This value is the range in which the setpoints are allowed to deviate in percent or units. The Relay Setpoints ON and OFF values are randomly varied inside the range to ensure that the material level does not consistently stop at the same point.
![Siemens 200 — P136 Wall Cling Reduction [MR 200] diagram (page 153)](/cdn/reader/82/829c15aaebfbf17c854075ac1e0dbc01f9b82abbfd8c8576c5c8e04c8c601513/images/fig-153-2.webp)
Puts pumps into groups for multiple pump rotations on one transducer.
![Siemens 200 — P137 Pump Group [MR 200] diagram (page 153)](/cdn/reader/82/829c15aaebfbf17c854075ac1e0dbc01f9b82abbfd8c8576c5c8e04c8c601513/images/fig-153-3.webp)
Alters • P111 Relay Control Function when P111=52 (Alternate duty assist) or 53 (Alternate duty backup)
This feature groups pumps (relay points 1 - 6) into groups 1 or 2. It is applied to pump rotation and occurs independently within each group.
![Siemens 200 — P137 Pump Group [MR 200] diagram (page 153)](/cdn/reader/82/829c15aaebfbf17c854075ac1e0dbc01f9b82abbfd8c8576c5c8e04c8c601513/images/fig-153-4.webp)

Please note that these parameters apply to the MultiRanger 200 only.
Use this feature to control an electrically operated flush valve on a pump to divert some pump output back into the wet well to stir up sediment.
Enter the MultiRanger relay number of the pump with the flush valve. The activation of this pump relay drives the usage of the flush system. Both P172 Flush Interval and P171 Flush Cycles are based on the operation of this relay and control any relay set to P111 = 64, Flush Valve.
The indexed relay is the one that controls the flush device. The value is the pump relay that is watched by the flush system. Enter the pump relay value into the parameter at the flush relay index.
If you need to watch pump Relay One to control a flush valve on Relay Two you would set P170[2]=1.
Picks the number of the pump relay which triggers the flushing device.
Related • P111 = 64, Flush Valve
Enter the MultiRanger relay number of the pump with the flush valve. The activation of this pump relay drives the usage of the flush system. Both P172 Flush Interval [MR 200] and P171 Flush Cycles [MR 200] are based on the operation of this relay and controls any relay set to P111 = 64, Flush Valve.
Primary Index · Global Relay · Values

Sets the number of pump cycles requiring flush control.
Related • P111 = 64, Flush Valve
P172 (Flush Interval) = 10 P171 (Flush Cycles) = 3
Sets the number of pump cycles before flush control is enabled.
Related • P111 = 64, Flush Valve
To start a new flush cycle every ten times the pumps are run, set this to 10.
The length of time for each flush cycle that the flush control is active.
Related • P111 = 64, Flush Valve

P200 mA Output Range
Determines the mA output range.
Primary Index mA output
1 0 to 20 mA
2 * 4 to 20 mA
3 20 to 0 mA
4 20 to 4 mA
Related • P911 mA Output Value
If either 1 or 2 is selected, the mA output is directly proportional to the mA Function. If either 3 or 4 is selected, then the output is inversely proportional. After setting P200, verify that P212 has a valid entry for mA Output Min Limit, as P212 is not changed automatically by a change to P200.
Alters the mA output/measurement relationship.
Please note that parameter P201 values vary between the MultiRanger 100 and the MultiRanger 200.
Primary Index mA output
value mA function Operation (P001)
Altered By • P001 Operation

Primary Index mA output
value mA function Operation (P001)
1 level level, differential, or average
4 volume level or space
Altered By • P001 Operation
Sets the input source from which the mA output is calculated.
Please note that parameter P202 values vary between the MultiRanger 100 and the MultiRanger 200. MultiRanger 100
Primary Index mA output
1 * Point 1
1:2 Average of readings from Point 1 and Point 2
Related • P201 mA Output Function
Primary Index mA output
Values 1 * Point 1
1:2 Average of readings from Point 1 and Point 2
Related • P201 mA Output Function

Enter the Point Number the mA output is based on. This value depends on whether mA function (P201) is set as transducer or mA input.
For a single point device, if P201 uses a transducer, parameter P202 can only be altered if P001 (Operation) is set for DPD or DPA.
P202 can be set to a particular point or range of points. When set to a range of points, the mA output will be the average of the readings from all in service transducers in the range. Out of service transducers will be ignored.
If P202 mA Output Allocation End contains the value 0, then only 1 transducer has been assigned to the mA output.
Displays current mA output value for the Point Number.
This displays as an Auxiliary Reading when key is pressed in the RUN mode and does not include adjustments made using Trim features (P214 / P215).
P203 holds the mA value associated with a transducer. This value is used for run mode display of the mA output. When only one mA output is associated with a transducer, this parameter holds the value of that mA output. When more than one transducer is associated with a mA output (set via P202), P203 holds the mA output calculated from the resulting average of the associated transducer readings. When more than one mA output is associated with a transducer, P203 holds the value of the first mA output associated with the transducer.
Primary Index Level

Use these features to reference the minimum and/or maximum mA output to any point in the measurement range.
Settings Action
Level, Space, or Distance Enter the material level in Units (P005) or percent of Span (P007) as referenced to Empty (P006).
Volume [MR 200] Enter the volume in Max Volume (P051) units or as a per- cent of Max Volume.
Flow [MR 200] Enter the flowrate in OCM Max Flow (P604) units or as a percent of OCM Max Flow.
Head [MR 200] Enter the head in level units (P005) or percent of Max Head (P603).
Volume Rate [MR 200] Enter the volume rate in volume/min. Ensure the % sym- bol is displayed before attempting to enter a % value.
mA input or Communications Input Not Applicable
Sets the process level corresponding to the 0 or 4mA value.
Primary Index mA output
Related • P211 20 mA Output Setpoint
Enter the value (in applicable units or %) to correspond to 0 or 4 mA.
Sets the process level that corresponds to the 20 mA value.
Primary Index mA output
Related • P210 0/4 mA Output Setpoint
Enter the value (in applicable units or %) to correspond to 20 mA.
Use these features to adjust the minimum and/or maximum mA output values, which should suit the input limit requirements of the external device.

Sets the minimum mA output value (in mA) to be produced.
Primary Index mA output
Sets the maximum mA output value (in mA) to be produced.
Primary Index mA output
Related • P200 mA Output Range / P212 mA Output Min LImit
This does not affect the P203 value shown, and is used when recalibration of an external device is impractical.
To adjust the value so that the device correctly indicates 4.00 (when P214 is accessed) or 20.00 mA (when P215 is accessed):
ENTER . The ammeter should show a value near 4 mA. 3. Enter the exact value displayed on the ammeter into P214 (Index 1 or 2). 4. The ammeter should then read exactly 4.00 mA. 5. Repeat steps 1 to 4 to set P215, using 20 mA as the desired value.
Calibrates the 4 mA output.
Primary Index mA output
Related • P215 20 mA Output Trim
Adjust this value so the device indicates 4.000 mA when P214 is accessed.
Values · Values
![Page 161 of the Siemens 200 user manual — mA Input (P250 to P260) [MR 200]](/cdn/reader/82/829c15aaebfbf17c854075ac1e0dbc01f9b82abbfd8c8576c5c8e04c8c601513/images/page-161.webp?w=400)
Calibrates the 20 mA output.
Primary Index mA output
Related • P214 4mA Output Trim
Adjust this value so the device indicates 20.00 mA when P215 is accessed.
Please note that this parameter applies to the MultiRanger 200 only.
Use for failsafe operation, independent of the Failsafe Material Level (P071).
Primary Index mA output
OFF * mA output responds to Failsafe Material Level (P071).
HOLd last known value is held until normal operation resumes
LO produce the Empty mA output immediately
HI produce the Span mA output immediately
Related • P201 mA Output Function
Or, to produce an mA output at a specific value, enter the value required. This is used only if mA output is allocated to a transducer (P201 = 1 to 7).
Please note that these parameters apply to the MultiRanger 200 only.
Values
![Page 162 of the Siemens 200 user manual — P252 20 mA Input Level [MR 200]](/cdn/reader/82/829c15aaebfbf17c854075ac1e0dbc01f9b82abbfd8c8576c5c8e04c8c601513/images/page-162.webp?w=400)
Shows the mA input range of the connected mA device.
1 0 to 20 mA
2 * 4 to 20 mA
Ensure this range corresponds to the output range of the external device. All level measurements will equate % of Span with the % of the mA range.
Shows the process level corresponding to the 0 or 4 mA value.
When using an external mA signal to determine level, the input range must be scaled to give accurate results.
Shows the process level corresponding to the 20 mA value.
Input range is scaled for accuracy if an external mA signal calculates level.

Shows the time constant used in the mA input filter to dampen signal fluctuations.
This number in seconds is used in the damping calculations. Larger values damp more than smaller values and 0 disables the signal filter.
Shows the resulting level value after scaling.
This parameter is calculated from the input mA signal.
Shows the raw mA input supplied by the external device.
Primary Index mA input
Discrete inputs can be used for the following:
Use the parameters listed above to have discrete inputs modify the unit’s operation.
Use the following parameters to configure the discrete input itself.
See also the Pump Control section on page 58 for a description of the MultiRanger’s pump control algorithms, including how the discrete inputs alter its operation.

Sets how discrete signals are interpreted by the MultiRanger.
Primary Index Discrete input
2 * Normally Open – 0 (DI open), 1 (DI closed)
3 Normally Closed – 0 (DI closed), 1 (DI open)
Related • Pump Control section
Shows the current value of the discrete input after any scaling is applied.
Primary Index Discrete input
0 (DI open), 1 (DI closed) Normally Open
0 (DI closed), 1 (DI open) Normally Closed
Readings are updated continuously even in PROGRAM mode. The value signals a level override event.
All records can be reset by pressing the CLEAR keys.
These features display the high and/or low temperatures in °C. When a parameter relating to a TS-3 Temperature Sensor is accessed, the Point Type display changes to the TS-3 symbol .
If the unit is powered up without a temperature sensor connected, the value –50°C is displayed. This information can help trace problems with both built in and external temperature sensors.

Shows the highest temperature encountered, as measured by the temperature sensor in the transducer (if applicable).
Related • P301 Temperature, Transducer Min
Press CLEAR keys to reset the log after a short circuit on the transducer wiring.
View the lowest temperature encountered, as measured by the temperature sensor in the transducer (if applicable).
Related • P300 Temperature, Transducer Max
Press CLEAR keys to reset the log after an open circuit on the transducer wiring.
View the highest temperature encountered, as measured by the TS-3 Temperature Sensor (if applicable).
Related • P303 Temperature, Sensor Min
Press CLEAR keys to reset the log after a short circuit on the transducer wiring.

Shows the lowest temperature encountered, as measured by the TS-3 Temperature Sensor (if applicable).
Related • P302 Temperature, Sensor Max
Press CLEAR keys to reset the log after an open circuit on the transducer wiring.
This identifies the occurrence of the record high and low level readings. Press CLEAR
keys to reset these values once the installation is working correctly.
Shows the highest Reading calculated (in normal Reading units or %).
Related • P305 Reading Min
Shows the lowest Reading calculated (in normal Reading units or %).
Related • P304 Reading Max
These features to identify pump usage and if the associated Relay Function (P111) is set for any pump control feature. The value is that of the pump connected to the associated terminals.
Enter a value to set the current record to that value. Use this if a pump is added with a known number of hours logged, or the value can be reset to zero 0 after maintenance.
![Page 167 of the Siemens 200 user manual — P312 Pump Run Ons [MR 200]](/cdn/reader/82/829c15aaebfbf17c854075ac1e0dbc01f9b82abbfd8c8576c5c8e04c8c601513/images/page-167.webp?w=400)
Displays the amount of time in minutes since a relay was last activated.
Related • Relay Function (P111) set for any pump control feature
Parameter measures the length of time since a relay was asserted, most often to determine how long a pump has been running. Alternatively, it can monitor a relay to show how long it has been in a state of alarm. It resets every time the relay is activated.
View or reset the accumulate ON time for the displayed Relay Number.
Related • Relay Function (P111) set for any pump control feature
Value is displayed with a floating decimal point (the more figures displayed before the decimal, the fewer displayed after). It is the value displayed when key is pressed in the RUN mode.
View or reset the accumulated number of times the Relay Number has been ON.
Related • Relay Function (P111) set for any pump control feature
This value is displayed when key is pressed and held for five seconds in RUN mode.
View or reset the accumulated number of times the displayed Relay Number has been held ON via Run On Interval (P130).
Please note that this parameter applies to the MultiRanger 200 only.
Related • Relay Function (P111) set for any pump control feature

Please note that these parameters apply to the MultiRanger 200 only.
These features are enabled if Operation is set for OCM (P001 = 6), or an OCM device is defined (P600 ≠ 0). Use them to identify the occurrence of the record high and low flow rates as displayed in OCM Max Flow (P604) units, or as a percent of OCM Max Flow. Press CLEAR keys to reset values once the installation is working correctly.
View the highest flow rate calculated (in units or %).
Related • P604 Maximum Flow
View the lowest flow rate calculated (in units or %).
Related • P604 Maximum Flow
Please note that these parameters apply to the MultiRanger 200 only.
Use these features to view, reset, or preset the eight-digit display totalizer when Operation is set for OCM or Pumped Volume (P001 = 6 or 7). The eight-digit totalizer is divided into two groups of four digits. The four least significant totalizer digits are stored in P322, and the four most significant digits are stored in P323. Adjust these values separately to set a new total.
P323 = 0017 P322 = 6.294 Totalizer Display = 00176.294

Totalizer units are dependent upon programming. Enter zero 0 (if required) to reset the totalizer to zero. Alternatively, enter any other (applicable) value, to preset the totalizer to the necessary value.
View and/or alter the four least significant digits of the totalizer value.
View and/or alter the four most significant digits of the totalizer value.
These features can record up to ten Echo profiles, initiated manually (P330), or automatically (P331 et al). If ten Profiles are already saved, addresses 1 through 10 are filled, the oldest automatically initiated record is overwritten. Manually initiated records are not automatically overwritten. All records are automatically deleted in the event of a power interruption.
When a record is displayed, results are based on current programming (which may have been altered since the record was saved). This permits the effect on the echo profile to be observed when changing an echo parameter.

Records profiles for later viewing.
Primary Index Echo profile
A1 automatically recorded profile from Transducer One
A2 automatically recorded profile from Transducer Two
U1 manually recorded profile from Transducer One
U2 manually recorded profile from Transducer Two
Press TRANSDUCER to fire the transducer and record the echo profile into the internal buffer for display.
To save a manual record
Press ENTER to copy the echo profile record in the buffer and save it in the selected address in the record library. The parameter value field displays the new record information.
Press key to enter display auxiliary mode and then:
Press CLEAR and then ENTER to delete the echo profile record in the selected address. The value returns to -.
no record · Values

Use to enable/disable the Auto Profile Record function.
Specifies the Transducer Point Number for which Auto Profile Records are saved.
Please note that Differential and Average operation is a MultiRanger 200 feature.
1 * Transducer One
Altered By • P001 Operation = 4 or 5
This feature is preset to Point Number 1. (Alteration is only required if differential or average Operation [P001 = 4 or 5] is selected.)
Enter the time to elapse after an Auto Profile Record is saved before another Auto Profile Record can be saved (subject to all other restrictions).

Use Auto Record ON Setpoint (P334) and Auto Record OFF Setpoint (P335) to define the boundaries within which the level must be, for the resultant Echo Profile to be considered for an Auto Profile Record.
If is displayed for either P334 or P335, Auto Profile Records are saved regardless of current level (subject to all other restrictions).
Enter the level value in Units (P005) or percent of Span (P007) as referenced to Empty (P006).
Enter the critical level which, in conjunction with Auto Record OFF Setpoint, defines the boundaries for Auto Profile Records to be saved.
Enter the critical level which, in conjunction with Auto Record ON Setpoint, defines the boundaries for Auto Profile Records to be saved.

Use this feature to restrict Auto Profile Records from being saved unless the level is rising, falling or either.
0 * Auto Profile Record on filling or emptying
1 Auto Profile Record on filling only
2 Auto Profile Record on emptying only
If the level changes at a rate in excess of the corresponding Filling / Emptying Indicator (P702 / P703) values, the Echo Profile is saved subject to this and other Auto Profile Record restrictions.
Limits Auto Profile Records from being saved unless extended LOE occurs.
If the LOE condition exceeds the period entered, the Echo Profile is saved. When set for 0 LOE is not required for an Auto Profile Record to be saved.

P340 Date of Manufacture
View the date of manufacture of this MultiRanger unit.
View the number of days this MultiRanger has been in operation.
The RUN Time value is updated once a day, and cannot be reset. However, in the event of a power interruption, the counter won’t advance. Therefore, a unit that is powered down on a regular basis will not have an accurate value.
The number of times power has been applied since the Date Of Manufacture.
Primary Index Global · Primary Index Global · Primary Index Global

Open Channel Monitoring (OCM) operation is a MultiRanger 200 feature only.
If the MultiRanger is used to monitor open channel flow, alter the following parameters as required and run a calibration as described in P621.
The MultiRanger measures head as referenced to Empty (P006) or OCM Zero Offset (P605), when Operation is set for OCM (P001 = 6). Flowrate, based on head (at the point of measure specified by the Primary Measuring Device fabricator) is also calculated and displayed on the LCD.
Some Primary Measuring Devices require a longer Range Extension (P801) to avoid entering the LOE failure state if the water level falls below the zero point of the Primary Measuring Device. See P801 Range Extension on page 200 for more information.
The type of primary measuring device (PMD) used.
0 * off (no calculation)
1 Exponential (see P601)
2 Palmer-Bowlus Flume (see P602)
3 H-Flume (see P602)
4 Universal Linear Flow Calculation (see P610, P611)
5 Universal Curved Flow Calculation (see P610, P611)
6 BS-3680/ISO 4359 Rectangular Flume (see P602)
7 BS-3680/ISO 1438/1 Thin Plate V-Notch Weir (see P602)
Altered By • P001 Operation

The MultiRanger is pre-programmed for common PMD flow calculations. If your PMD is not listed, select the appropriate Universal Flow Calculation.
Associated parameters Max Head (P603), Max Flow (P604), and Min Head (P605) may be scroll accessed. If Operation is not set for OCM (P001 = 6), this value is preset to 0. If Operation is set for OCM, it is preset to 1.
The Exponent for the flow calculation formula.
Altered By • P600 Primary Measuring Device
Use this parameter if the Primary Measuring Device (P600) is set to 1 (exponential). It creates an exponential curve with end points set by Max Head (P603) and Zero Head (P604) and with the curve based on the specified exponent.
If P601 = 1, the flow characteristic is a straight line
Max. Flow
(P604)
The exponential equation is
Max. Head
Use the exponent specified by the PMD manufacturer, if available, or the sample value given below.

Suppressed Rectangular Weir 1.50
Cipolletti Weir 1.50
Venturi Flume 1.50
Parshall Flume 1.22 to 1.607
Leopold Lagco 1.547
V-Notch Weir 2.50
The dimensions of the Primary Measuring Device (PMD).
Global Transducer and Dimension
ISO 4359
Index Values for Supported PMDs
Altered By • P600 Primary Measuring Device
Use this parameter if the Primary Measuring Device is directly supported (P600=2,3,6,7). The dimensions required for each PMD vary.
For more information on PMD, see page 73.

The level value associated with Maximum Flow, in Units (P005).
This represents the highest head level supported by the PMD and works in conjunction with Maximum Flow (P604) to define the highest point in the exponential curve. Use it when the Primary Measuring Device (PMD) requires a maximum head and flow reference point. This would include Exponential, Palmer Bowlus Flume, H-Flume, and Universal breakpoints.
The maximum flowrate associated with Maximum Head (P603).
Altered By • P600 Primary Measuring Device
This represents the flow at the highest head level supported by the PMD. and works in conjunction with Maximum Head (P603) to define the highest point in the exponential curve. Use it when the Primary Measuring Device (PMD) requires a maximum head and flow reference point. This would include Exponential, Palmer Bowlus Flume, H-Flume, and Universal breakpoints.
Also use this parameter with Time Units (P606) to define the flowrate units. The limitation of four digits is for the LCD only, and the flowrate value is available with greater precision through communications.

Conditions Enter
The distance above Empty (P006) in Units (P005) representing zero head (and zero flow).
This feature can be used for most weirs and some flumes (e.g. Palmer Bowlus) where the zero reference is at a higher elevation than the channel bottom.
Material level at zero flow

Determines the units used to display current flow and logging flow values.
Altered By • P608 Flowrate Units
This is used when the Primary Measuring Device is Ratiometric (P608=0).
Conditions Enter
The maximum number of decimal places to be displayed.
0 no digits after the decimal point
1 1 digit after the decimal point
2 2 digits after the decimal point
3 3 digits after the decimal point
Altered By • P060 Decimal Position
In RUN mode, the number of decimal places displayed is automatically adjusted (if necessary) to prevent the number of Flowrate digits from exceeding display capabilities.
The maximum number of head decimal places is controlled by Decimal Position (P060).

The volume units used to display total flow.
0 * Ratiometric calculation (units defined by P604)
1 litres / second
2 cubic metres / hour
3 cubic metres / day
4 cubic feet / second
5 gallons / minute – Imperial
6 million gallons / day – Imperial
7 gallons / minute – U.S.
8 million gallons / day – U.S.
Alters • P606 Time Units
Altered By • P600 Primary Measuring Device
Related • P608 Flowrate Units
This parameter is enabled only if the primary measuring device (PMD) supports absolute calculations (P600=6,7). For absolute PMDs (P600=6,7) volume units can be specified using this parameter. If needed, absolute PMDs can still use ratiometric (P608=0) to accommodate other units.

The head breakpoints for which flowrate is known. See “Accessing a Secondary Index” on page 117 for more information..
Related • P611 Breakpoint Flowrates
The values in the Span for which flowrates are known. See Universal Calculation Support on page 83 for how to specify universal flows.
The flowrate corresponding to each Head Breakpoint entered.
Related • P610 Head Breakpoints
These are the flowrates for the related breakpoints. See Universal Calculation Support on page 83 for how to specify universal flows.
Max. Flow
Max. Head

Eliminates totalizer activity for flows at or below the cutoff value.
Preset = 5.000 %, or equivalent units
Altered By • P005 Units
Related • P007 Span
Use this to enter the minimum head in units (P005) or as a percent of span.
Calibrates Zero Head (P605) based on actual head measurements.
Use this parameter when the reported head is consistently high or low by a fixed amount.
The deviation between the entered Empty (P006) value and the calibrated Empty value is stored in Offset Correction (P652). Alternatively, the Empty parameter (P006) can be corrected directly.
![Page 184 of the Siemens 200 user manual — P622 Inflow / Discharge Adjust [MR 200]](/cdn/reader/82/829c15aaebfbf17c854075ac1e0dbc01f9b82abbfd8c8576c5c8e04c8c601513/images/page-184.webp?w=400)
The Pumped Volume Totalizer is a MultiRanger 200 feature only.
If the eight-digit totalizer display or a remote totalizer contact closure are desired, alter the following parameters.
The method used to calculate the volume pumped, for pumped total Operation (P001 = 7).
Primary Index Single Point Model Dual Point Model Global Transducer
1 = inflow * / pump cycle When the pump is off, the MultiRanger estimates the volume of inflow by recording the rate at which the liquid level changes. When the pump is operating, the estimated inflow volume is added to the pumped volume total. When the pump stops, the pumped volume of the previous pump cycle is added to the total volume pumped in the totalizer. 2 = inflow * ignored Inflow is assumed to be 0 while pumps are running. 3 = inflow * / rate (preset) Volume pumped is adjusted for inflow. Inflow rate is estimated by assuming that the rate calculated (P708) just prior to the start of the pump cycle remained constant during the pump cycle. Inflow rate is averaged using rate filter (P704), rate update time (P705) and rate update distance (P706) to control how the average rate is calculated.
Values · Related · * or discharge

The Totalizer features apply to MultiRanger 200 only.
Use this feature if the LCD Total increments by too large (or too small) an amount.
-3 .001
Related • LCD Totalizer (P322 and P323)
Enter the factor (powers of 10 only) by which actual volume is divided, prior to display on the LCD. Use a value such that the eight-digit totalizer doesn’t roll over between readings.
For an LCD Total display in 1000s of volume units, enter 3.

Enter the maximum number of decimal places to be displayed.
0 no digits after the decimal point
1 1 digit after the decimal point
2 * 2 digits after the decimal point
3 3 digits after the decimal point
Related • LCD Totalizer (P322 and P323)
In RUN mode, the number of decimal places displayed is not automatically adjusted. When the LCD Total value is so large as to exceed display capabilities, the total rolls over to 0 and continues incrementing.
Use this feature if the remote totalizer (device connected to the relay set for totalizer operation [relay Function, P111 = 40]), updates too slowly or rapidly.
-3 .001
Parameter is relevant only if Operation is set to OCM or Pumped Volume (P001 = 6 or 7). The relays on the MultiRanger have a maximum frequency of 2.5 Hz.

Enter the factor (powers of 10 only) by which actual volume is divided, prior to Remote Totalizer count increment.
For a Remote Totalizer update by 1000s of volume units, enter 3.
Use this feature in conjunction with Flow Sampler Exponent (P642) to establish the number of flow units required to increment the Flow Sampler (device connected to the MultiRanger relay set for the flow sampler operation Relay Function, P111 = 41).
Preset = 1.000
This parameter is relevant only if Operation is set to OCM (P001 = 6). Enter the mantissa (Y) for the exponent (Z) in the formula:
Flow Sampler Increment = Y x 10Z Flow units.
Use this feature in conjunction with Flow Sampler Mantissa (P641) to establish the number of flow units required to increment the Flow Sampler (device connected to the MultiRanger relay set for the flow sampler operation Relay Function, P111 = 41).

This parameter is relevant only if Operation is set to OCM (P001 = 6).
Flow Sampler Increment = Y x 10Z Flow units.
Use this feature (if desired) to adjust the minimum contact closure duration of a relay set as a totalizer, flow sampler, time (control), or aeration (P111 = 40, 41, 60 or 62)
Preset = 0.2 (sec)
Related • P111 Relay Control Function
Enter minimum contact closure duration (in seconds) required by the device connected.
For the flow sampler function this value is used for both the ON time of the relay and the OFF time between contacts.
P645 OFF time P645 on time relay contact closed
There are two types of calibration possible:
Offset: Adjusts the measurements by a fixed amount.
Sound Velocity: Adjusts speed of sound and changes the measurement calculations.
Do Offset calibration at any steady level unless a Sound Velocity calibration is also done. If both calibrations are done then do Offset at a known high level and Sound Velocity at a known low level.
Primary Index Global · Values · relay contact open

Calibrates Empty (P006) if the reported level is consistently high or low by a fixed amount.
Begin with a steady level.
Changes the speed of sound constant.
For best results, calibrate with the level at a known value near empty.

Ensure a steady level at some low value (P653 and P654 adjusted accordingly)
The value altered when an Offset Calibration is performed.
Related • P650 Offset Calibration
Alternatively, if the amount of Offset Correction required is known, enter the amount to be added to the Reading before display.
The value adjusted based on the Sound Velocity at 20 °C (P654) vs. Temperature (P664) characteristics of air.
Alternatively, enter the current sound velocity (if known), or perform a Sound Velocity Calibration (P651). The units used are m/s if P005 = 1, 2, or 3 (ft/s if P005 = 4 or 5).

This value is used to automatically calculate Sound Velocity (P653).
After performing a Sound Velocity Calibration, check this value to verify the acoustic beam atmosphere is air (344.1 m/s or 1129 ft/s).
Alternatively, if the acoustic beam atmosphere sound velocity at 20°C (68 °F) is known, and the sound velocity vs. temperature characteristics are similar to that of air, enter the sound velocity.
The units used are m/s if P005 = 1, 2, or 3 (or ft/s if P005 = 4 or 5).
P660 Temp Source
Source of the temperature reading used to adjust the speed of sound.
4 TS-3 Temperature Sensor
5 Average (TS-3 and transducer)
Alters • P664 Temperature
The MultiRanger uses the TS-3 temperature sensor assigned to the transducer. If one is not connected, the ultrasonic/temperature transducer is used. If the transducer does not have an internal temperature sensor, the Temp Fixed (P661) value is used.
If the acoustic beam atmosphere temperature varies with distance from the transducer, connect a TS-3 Temperature Sensor and ultrasonic / temperature transducers, and select average.

In gasses other than air, the temperature variation may not correspond with the speed of sound variation. In these cases turn off the temperature sensor and use a fixed temperature.
Use this feature if a temperature sensing device is not used.
Enter the temperature (in °C) of the atmosphere within the transducer acoustic beam. If the temperature varies with distance from the transducer, enter the average temperature.
This feature may only be used for differential or average Operation (P001 = 4 or 5).
1 * Transducer One
1:2 Transducer One and Two average [MR200]
As preset, the temperature measurements of Ultrasonic / Temperature Transducer One and Two are allocated to Points 1 and 2 respectively.
Use this feature if the temperature measurement from both transducers should be identical, but one is located close to a radiant heat source. Allocate the temperature measurement of the other transducer to both transducer Point Numbers.
Enter the number of the Transducer whose temperature measurement will be used for the distance calculation of the Point Number displayed. When both transducers are allocated to a Point Number, the temperature measurements from each are averaged.

View the transducer temperature in °C.
Altered By • P660 Temp Source
Value is displayed when is pressed in RUN mode (see Readings in Run Mode on page 23).
If Temp Source (P660) is set to any value other than Fixed Temp, the value displayed is the temperature measured. If Temp Source is set to Fixed Temp, the P661 value is displayed.
These parameters determine how material level changes are reported.
Adjusts the MultiRanger response to increases in the actual material level (or advance to a higher Failsafe Material Level, P071).
Altered by • P003 Maximum Process Speed
Enter a value slightly greater than the maximum vessel filling rate. This value, in Units (P005) or % of Span (P007) per minute, is automatically altered when Maximum Process Speed (P003) is altered.

Adjusts the MultiRanger response to decreases in the actual material level (or advance to a lower Failsafe Material Level, P071).
Altered by • P003 Maximum Process Speed
Enter a value slightly greater than the maximum vessel emptying rate. This value, in Units (P005) or % of Span (P007) per minute, is automatically altered when Maximum Process Speed (P003) is altered.
P702 Filling Indicator
The fill rate required to activate the LCD Filling indicator ( ).
Altered by • P003 Maximum Process Speed
This value (in Units (P005) or % of Span (P007) per minute) is automatically set to 1/10 of the Max Fill Rate (P700).

The empty rate required to activate the LCD Emptying indicator ( ).
Altered by • P003 Maximum Process Speed
This value (in Units (P005) or % of Span (P007) per minute) is automatically set to 1/10 of the Max Empty Rate (P701).
Damps Rate Value (P707) fluctuations.
0 Rate display not required
1 Continuously filtered and updated
2 1 minute or 50 mm (2 in)
3 5 minutes or 100 mm (3.9 in)
4 10 minutes or 300 mm (11.8 in)
5 10 minutes or 1000 mm (39.4 in)
Alters • P707 Rate Value
Altered by • P003 Maximum Process Speed
Related • P705 Rate Update Time/ P706 Rate Update Distance
Enter the time or distance interval over which the Rate Value is to be calculated before the display updates.
This is automatically altered along with Maximum Process Speed (P003).
This value automatically alters the Rate Update Time (P705) and / or Rate Update Distance (P706). Alternatively, these parameter values may be altered independently.

The time period (in seconds) over which the material level rate of change is averaged before Rate Value update.
Related • P707 Rate Value
The material level change (in metres) to initiate a Rate Value update.
Related • P707 Rate Value
The rate of material level change (in Units (P005) or % of Span (P007) per minute).
Altered By • P704 Rate Filter
A negative rate indicates the vessel is emptying.
This is the value displayed when is pressed in the RUN mode as described in the Readings in RUN Mode chart on page 23.

The rate of change of volume in percent of maximum volume per minute.
Please note that this feature applies to the MultiRanger 200 only.
Related • P622 Inflow / Discharge Adjust
This value is used internally to calculate inflow in pumped volume applications (P622=3).
Press READING to toggle between percent and volume.
P710 Fuzz Filter
Use this to stabilize the reported level, due to level fluctuations (such as a rippling or splashing liquid surface) within the Echo Lock Window (P713).
Altered by • P003 Maximum Process Speed
This value (in % of Span, P007) is automatically altered when Maximum Process Speed (P003) is altered. The higher the value entered, the greater the fluctuation stabilized.

Use this feature to select the measurement verification process.
2 * Material agitator
If a material agitator (mixer) is used in the vessel monitored, set Echo Lock for maximum verification or material agitator to avoid agitator blade detection. Ensure the agitator is always ON while the MultiRanger is monitoring the vessel to avoid stationary blade detection.
When set for max verification or material agitator, a new measurement outside of the Echo Lock Window (P713) must meet the sampling criterion (P712).
For total lock, Echo Lock Window (P713) is preset to zero 0. The MultiRanger continuously searches for the best echo according to the algorithm chosen (P820). If the selected echo is within the window, the window is then centered about the echo. If not, the window widens with each successive shot until the selected echo is within the window. The window then returns to its normal width.
When Echo Lock is OFF , the MultiRanger responds immediately to a new measurement as restricted by the Max Fill / Empty Rate (P700 / P701); however, measurement reliability is affected.

The sampling criterion sets the number of consecutive echoes appearing above or below the echo currently locked onto, that must occur before the measurements are validated as the new reading (for Echo Lock P711 values: 1 or 2).
x = the number of above echoes
y = the number of below echoes
Related • P711 Echo Lock
1, max verification 5:5
2, material agitator 5:2
Adjusts the size of the Echo Lock Window.
Altered by • P003 Maximum Process Speed

The Echo Lock Window is a distance window (units P005) centred on the echo and used to derive the Reading. When a new measurement is in the window, it is re-centred and the new Reading calculated. Otherwise, the new measurement is verified by Echo Lock (P711) before the reading is updated.
When 0 is entered the window is automatically calculated after each measurement. For slower P003 Maximum Process Speed values the window is narrow, for faster P003 values the window becomes wider.
P726 Level System Sync
Enables the System Sync on the terminal block.
1 * synchronize level monitors
Use this if another level measurement system is mounted nearby, and they are wired together on the Sync terminal.
The delay, in seconds, between measurements from transducer points (dual point model only).
Altered by • P003 Maximum Process Speed
Related • P001 Operation
This feature may only be used to adjust the delay before the next point is scanned. Enter the amount of delay in seconds. This value is automatically altered when Maximum Process Speed (P003) is altered.

The delay, in seconds, between transducer shots.
Use this if transient acoustic noise within the vessel is causing measurement difficulties due to echoes from one shot being received on the next. If more than one ultrasonic unit is installed for redundancy, this value should be 0.
View the elapsed time (in seconds) since the point displayed was last scanned.
Related • P001 Operation
This may be viewed as an Auxiliary Reading in the RUN mode.
P730 Auxiliary Reading
Use this feature to display operator selected Auxiliary Readings temporarily or indefinitely (as desired).
Select OFF to display Auxiliary Readings temporarily. Select HOLd to display Auxiliary Readings until another Auxiliary Reading is selected or programming mode is entered. See the Hand Programmer section on page 27 for RUN mode auxiliary readings.

If necessary, enter the Parameter Number to default in the Auxiliary Reading display. That value will show in the auxiliary reading area by default. Other values are available but will reset to the parameter defined here.
Enter the Parameter Number whose value is to be displayed in the Auxiliary Reading field
is pressed in the RUN mode. See the Hand Programmer section on page 27 for RUN mode auxiliary readings.
P732 Display Delay
Adjusts the Point Number display scroll speed.
Use this feature to adjust the delay before the display advances to the next Point Number. Display scrolling is independent from transducer scanning.
Use this feature to select the parameter scroll access option desired.
0 Off to scroll to all parameters (P001 to P999)
1 * Smart for Quick Start, altered, and tagged parameters
2 Tagged to scroll to operator tagged parameters only
Press READING and to tag / untag any accessed parameter. is displayed to indicate the parameter accessed is tagged.

Controls the LCD backlighting.
The backlight can be forced on or off, or be controlled by a programmer, in which case it will turn OFF 30 seconds after the last key is pressed.
The reading shown on the primary reading display when in RUN mode.
1 * Default reading (P920) based on operation (P001)
2 LCD totalizer (P322, P323)
3 Automatically toggle between 1 and 2
When this value indicates TOGGLE, then both readings (default and totalizer) are shown in the time specified in display delay (P732).
The maximum time allowed between receiving a request and transmitting the response.
If the maximum time is exceeded, no response will be transmitted, and the action required may not be completed.

These parameters are reserved for optional SmartLinx communications cards and vary by card. Refer to the SmartLinx documentation to determine if any of them are used.
The MultiRanger communication ports are configured by a series of parameters that are indexed by port. See the Communications section on page 89 for a complete description of communications set-up.
Communication parameters are indexed to these communication ports, unless otherwise noted:
1 RS-232 port (RJ-11 modular telephone)
2 RS 485 port on terminal block
The communications protocol used between the MultiRanger and other devices.
Primary Index Communications Port
0 Communications port disabled
1 Siemens Milltronics Dolphin protocol
2 Modbus ASCII slave serial protocol
3 * Modbus RTU slave serial protocol (preset for ports 1 and 2)
The MultiRanger supports the internationally recognized Modbus standard in both ASCII and RTU formats. Other protocols are available with optional SmartLinx cards.
The unique identifier of the MultiRanger on the network.
Primary Index Communications Port
1 * Preset:
For devices connected with the Siemens Milltronics protocol this parameter is ignored. For devices connected with a serial Modbus slave protocol, this parameter is a number from 1-247. The network administrator must ensure that all devices on the network have unique addresses. Do not use the value 0 for Modbus communications as this is the broadcast address and is inappropriate for a slave device.
Values · Values

The communication rate with the master device.
Primary Index Communications Port
4.8 4800 baud
9.6 9600 baud
19.2 * 19,200 baud (preset for port 2)
115.2 * 115,200 baud (preset for port 1)
This specifies the rate of communication in Kbaud. Any value may be entered but only the values shown above are supported. The baud rate should reflect the speed of the connected hardware and protocol used.
The serial port parity.
Primary Index Communications Port
0 * No Parity
Ensure that the communications parameters are identical between the MultiRanger and all connected devices. For example, many modems default to N-8-1 which is No parity, 8 data bits, and 1 stop bit.
The number of data bits per character.
Primary Index Communications Port
8 * Modbus RTU
7 or 8 Modbus ASCII
7 or 8 Dolphin Plus

The number of bits between the data bits.
Primary Index Communications Port
1 * Preset:
Sets the MultiRanger to use an external modem..
Primary Index Communications Port
0 * No modem connected
Sets the time that the unit will keep the modem connected with no activity.
Primary Index Communications Port
0 * No timeout
To use this parameter, ensure that P778 (Modem Available) = 1. Ensure that the value is low enough to avoid unnecessary delays when an unexpected disconnect occurs but long enough to avoid timeout while you are still legitimately connected. This parameter value is ignored by the Modbus Master Drivers, as they automatically disconnect when done.
If the line is idle and the P779 Modem Inactivity Timeout expires, then the modem is directed to hang up the line. Ensure that P779 is set longer than the standard polling time of the connected master device. 0 disables the inactivity timer.

Determines where index information is stored for the parameter access area.
Altered By • P770 Port Protocol
The primary and secondary index values are encoded into the format words found between R46,000 and R46,999. Each format work corresponds with the R44,000 series number in the parameter access map. For example, the format register R46,111 corresponds to the parameter P111 and the value is stored in R44,111. If the Modbus protocol (P770 = 2 or 3) is not used this parameter is ignored.
The results of ongoing hardware tests in the communications circuitry.
PASS * No errors
FAIL Error occurred communicating with card; commu- nications should resume.
ERR1 No module installed, or module not supported; SmartLinx communications have been disabled
If FAIL or ERR1 is displayed in P790 (Hardware Error), go to P791 (Hardware Error Code) and P792 (Hardware Error Count) for information about the error.

Indicates the precise cause of Fail or ERR1 condition from P790.
0 * No error
8 No SmartLinx card installed
Error code; provide this code to your Siemens Milltronics representative for troubleshooting
Related • P790 Hardware Error
A count that increments by 1 each time Fail is reported in P790 (Hardware Error).
Error count; provide this number to your Siemens Milltronics repre- sentative for troubleshooting.
Related • P790 Hardware Error
This parameter is used to identify the module type when SmartLinx is used. If you are not using SmartLinx, this parameter is not functional. Please see the associated SmartLinx instruction manual for a full description of this parameter.
This parameter is used to identify the protocol when SmartLinx is used. If you are not using SmartLinx, this parameter is not functional. Please see the associated SmartLinx instruction manual for a full description of this parameter.

Enables the read/write access to parameters via remote communications.
Protocol (Index 1 controls the Modbus Master (RS-485 or RS-232); Index 2 controls the Fieldbus Master (PROFIBUS DP, DeviceNet, or Allen Bradley Remote I/O)
2 Restricted Access – read only except for P799 which is read/write
P800 Near Blanking
The space near the transducer face which cannot be measured.
0.450m (XCT-8, XCT-12)
Use this feature if the surface is reported to be near the transducer face but is in fact much further away. Extend this value when changing transducer location, mounting, or aiming.
Please note that changing the Near Blanking cannot correct measurement problems. Ensure that Span (P007) < Empty (P006) minus Near Blanking (P800)

Allows the material level to fall below the Empty setting without reporting LOE.
This feature is useful in OCM applications where the Empty level is set to the bottom of the weir, and above the bottom of the channel, and should be used if the surface monitored can fall past the Empty (P006) level in normal operation. The value is added to Empty (P006) and can be greater than the range of the transducer. If the surface monitored can extend beyond Empty (P006), increase Range Extension (in Units (P005) or % of Span) such that Empty plus Range Extension is greater than the transducer face to furthest surface to be monitored distance. This is often the case with OCM when using weirs and some flumes.
Used when the transducer is expected to be submerged on occasion.
When a transducer with a submergence shield is submerged, the shield traps an air pocket that creates a special echo. The MultiRanger recognizes the echo and advances the reading to the highest level and operates displays and outputs accordingly. This feature is effective for when power is returned while the transducer is submerged.

Determines what type of ultrasonic shots are fired.
2 * Short and long
Increases MultiRanger response when the monitored surface is close to the transducer face. Select short and long to have short and long acoustic shots fired for each measurement, regardless of the transducer to surface distance. Select short to have only short shots fired if the Echo Confidence (P805) produced by a short shot exceeds the short Confidence Threshold (P804) and the monitored surface is always within the Short Shot Range (P852).
Determines which echoes are evaluated by software.
Related • P805 Echo Confidence
The short and long shot Confidence Thresholds are preset to 10 and 5 respectively. When Echo Confidence (P805) exceeds the Confidence Threshold, the echo is evaluated by Sonic Intelligence®. Values are entered as two numbers separated by a decimal point. The first number is the short shot confidence and the second number is the long shot confidence.

Displays the echo confidence of the measurement echo from the last shot.
x = short (0 to 99)
y = long (0 to 99)
Use this feature to monitor the effect of transducer aiming, location, and mechanical transducer / mounting isolation.
Both short and long shot Echo Confidence is displayed. (To display this value in the
auxiliary display while the unit is running, press for 4 seconds.)
Display Description x:-- short shot confidence value, (long shot not used). --:y long shot confidence value, (short shot not used). x:y short and long shot confidence values (both used). E transducer cable is open or short circuited. --:-- no shots were processed for Sonic Intelligence® evaluation.
Displays the strength (in dB above 1 uV RMS) of the echo which was selected as the measurement echo.
Displays the average and peak ambient noise (in dB above 1 uV RMS) being processed.
x = average (-99 to 99)
y = peak (-99 to 99)
The noise level is a combination of transient acoustic noise and electrical noise (induced into the transducer cable or receiving circuitry). See Noise Problems in the Troubleshooting Section on page 232.

The following parameters are for authorized Siemens Milltronics Service personnel or technicians familiar with Siemens Milltronics echo processing techniques.
The time (in ms) from the transmission of the pulse, to when it is processed.
Related • P816 Echo Time Raw
The time (in ms) from the transmit pulse to the processed echo.
Related • P815 Echo Time Filtered
Primary Index Transducer · Primary Index Transducer

Chooses the algorithm to generate the measured value from the profile.
Values 1 ALF = flat Area, Largest, and First average
2 A = flat Area only
3 L = flat Largest only
4 F = flat First only
5 AL = flat Area and Largest average
6 AF = flat Area and First average
7 LF = flat Largest and First average
8 * bLF = smooth Largest or First
9 bL = smooth Largest only
10 bF = smooth First only
12 tF = true First only
Use this to select the algorithm(s) the Sonic Intelligence® echo selection is based on. Use P805 Echo Confidence (page 202) to determine which algorithm gives the highest confidence under all level conditions. If the wrong echo is processed, observe the echo processing displays and select an alternate algorithm, either by entering the numeric value desired, or as below:
Primary Index · Global Transducer · Related

Dampens spikes in the echo profile to reduce false readings.
Use P821 if interference spikes are on the long shot Echo Profile display.
Filters out echoes of a specific width.
0 = OFF (preset), nearest acceptable value returned
Use this for transducer acoustic beam interference (e.g. ladder rungs). Enter the width of false echoes (in ms) to be removed from the long shot Echo Profile. When a value is keyed in, the nearest acceptable value is entered.

Smoothes jagged peaks in the echo profile.
0 = OFF (preset)
greater = wider, nearest acceptable value returned
Use this feature, when monitoring solids (P002 = 2), if the reported level fluctuates slightly, though the monitored surface is still. Enter the amount (in ms) of long shot Echo Profile smoothing required. When a value is keyed in, the nearest acceptable value is entered.
The point on the primary echo on which the measured value is based.
Use this feature if the reported material level fluctuates slightly, due to a variable rise in the leading edge of the true echo on the Echo Profile.
Enter the value (in percent of echo height) to ensure the Echo Lock Window intersects the Echo Profile at the sharpest rising portion of the Echo Profile representing the true echo. This value is preset to 50%.

The following parameters are for authorized Siemens Milltronics Service personnel or technicians familiar with Siemens Milltronics echo processing techniques.
Advanced TVT control applies to long shots only.
Selects the TVT Curve used.
1 * TVT Short Curved
2 TVT Short Flat
3 TVT Long Flat
4 TVT Long Smooth Front
5 TVT Long Smooth
Altered By • P002 Material
Select the TVT type which gives the highest confidence (P805) under all level conditions. Use this parameter with caution, and do not use TVT Slopes with the bF or bLF Algorithm (P820).
Turns the TVT Shaper ON or OFF.
Related • P832 TVT Shaper Adjust
Turn the TVT Shaper ON before using P832 and afterwards. Turn the TVT Shaper ON and OFF while monitoring the effect to pick up the true echo.

Allows manual adjustment of the TVT curve. Use this parameter in conjunction with Dolphin Plus PC Software.
Breakpoint Transducer and Breakpoint
Related • P831 TVT Shaper
Use this feature to bias the shape of the TVT curve to avoid selecting false echoes from fixed objects.
Adjustment to this parameter is best done while viewing the echo profile with Dolphin Plus. Refer to the Dolphin Plus online help for details.
The TVT curve is divided into 40 breakpoints, accessible by enabling the point number as the breakpoint index field. Each breakpoint is normalized to a value of 0, as displayed in the parameter value field. By changing the breakpoint value, up or down, the intensity of the bias applied to that breakpoint of the curve is respectively changed. By changing the value of adjacent breakpoints, the effective bias to the shaper can be broadened to suit the desired correction. In the case of multiple false echoes, shaping can be applied along different points of the curve. Shaping should be applied sparingly in order to avoid missing the true echo.
Use this feature to adjust the TVT Curve height to ignore false echoes (or pick up true echoes) near the start of the Echo Profile.
Enter the minimum TVT Curve start point (in dB above 1 uV RMS).
This feature should only be used if increased Near Blanking (P800) would extend farther than desired into the measurement range.

Use this feature in conjunction with TVT Start Min (P833) to ignore false echoes (or pick up true echoes) near the start of the Echo Profile.
Enter the time (in ms) for the TVT Curve to decrease from the TVT Start Min (P833) point to the TVT Curve baseline.
Enter the minimum slope (in dB/s) for the middle of the TVT Curve.
Use this feature to adjust the slope declination, and use it in conjunction with TVT Start Duration (when a long flat TVT Type is selected) to ensure the TVT Curve remains above the false echoes in the middle of the Echo Profile. Alternatively, if TVT Type is set for TVT Slopes (P830 = 6), preset is 2000.

Use P837 and P838 together, to set MultiRanger 100/200 to ignore false echoes. Use P838 to set the Auto TVT distance first.
If MultiRanger 100/200 displays a full level, or if the reading fluctuates between a false high level and a correct level, set P837 to elevate the TVT in this region and to de- sensitize the receiver from any ‘base noise’ caused by internal transducer reflections, nozzle echoes, or other vessel false echoes. Set P838 and then P837 (detailed instructions follow P838).
1 Use ’learned’ TVT. (See ’learned TVT curve’ in Display after Auto False Echo Suppression on page 211.)
Display before Auto False Echo Suppression
(or when P837 = 0)


Defines the range of Auto False-Echo Suppression (P837) to use for ignoring false echoes. (Units are defined in P005.)
Range (depends on model)
Default 1.000 m (3.28 ft)
Determine the actual distance from the transducer face to the material surface. Subtract 0.5 m from this distance, and enter the result.

TVT curve (learned) material level
Defines (in percent) how high the TVT curve is placed above the profile, relative to the largest echo. When MultiRanger 100/200 is located in the center of the vessel, lower this parameter to prevent multiple echo detections.
Values Range 0 to 100%

These parameters are for Siemens Milltronics service personnel only. P840 Short Shot Number
The number of short shots to be fired (and results averaged) per transmit pulse.
Enter the number of long shots to be fired (and results averaged) per transmit pulse.
Altered By • P003 Maximum Process Speed
This value is automatically altered by Maximum Process Speed (P003).

Adjust the short shot transmit pulse frequency (in kHz).
Altered By • P004 Transducer
This feature is automatically altered when Transducer (P004) is altered. P843 Long Shot Frequency
Adjust the long shot transmit pulse frequency (in kHz).
Altered By • P004 Transducer
This feature is automatically altered when Transducer (P004) is altered. P844 Short Shot Width
Adjust the width (in ms) of the short shot transmit pulse.
Altered By • P004 Transducer
This feature is automatically altered when Transducer (P004) is altered.

Adjust the width (in ms) of the long shot transmit pulse.
Altered By • P004 Transducer
This feature is automatically altered when Transducer (P004) is altered.
Use this feature to slant the echo evaluation in favour of the short shot echo when both short and long shots are evaluated (see Shot Mode, P803).
Enter the minimum echo strength (in dB above 1 uV) derived from a short shot to be considered for evaluation.

Enter the maximum distance in Units (P005) to be measured using short shot echoes.
Altered By • P004 Transducer
This feature is automatically altered when Transducer (P004) is altered.
Test Parameters are intended for use by Siemens Milltronics Service personnel.
View the EPROM Rev. #.
Press ENTER to activate the MultiRanger memory test.
PASS (memory test successful)

Press ENTER to put the CPU into an infinite loop to test the watchdog timer.
On successful completion (10 seconds) the RUN mode is entered and the MultiRanger is reset. Programming is kept and the unit responds as if there had been a power failure.
Press ENTER to activate the display test.
All LCD segments and symbols are temporarily displayed.
Press ENTER , then press each keypad key in the following sequence:
As each key is pressed, the associated keypad number is displayed. On successful test completion, PASS is displayed. FAIL is displayed if a key is pressed out of sequence or the programmer keypad malfunctions.
Press ENTER to supply repeated transmit pulses, at the frequency entered, to the transducer and / or view the transducer operating frequency (automatically altered by P004 Transducer) for the Point Number displayed.


Press ENTER to test the RS-232 port on the RJ-11.
An external device must be connected to the RS-232 port for this test. On successful completion, PASS is displayed, otherwise it is FAIL.

Press ENTER to cycle the scanner relay while firing the transmitter.
Use this parameter to ensure that both transducers are being stimulated.
Used to energize and de-energize relays directly.
Values 0 to 6
Related • P119 mA Relay Logic Test
Enter the relay number and then toggle between energized and de-energized, as required. Enter 0 to toggle all relays at once.
Applies only to relays with P119 = 0 (algorithm control). Use this parameter to confirm that relay contacts are opening and closing.
This feature is helpful when P119 does not give expected results even though programming was verified.
Access this parameter to display the current value of the mA output.
Primary Index mA output
Additionally, this feature may be used to enter a desired value. The mA output immediately assumes the value entered regardless of any restrictions programmed.
Use this feature to display the temperature in °C (as monitored by the connected transducer).
Err is displayed if the transducer is not equipped with an internal temperature sensor.
Primary Index Global · Primary Index Transducer

Access this parameter to display the temperature in °C (as monitored by the TS-3).
OPEn is displayed if a TS-3 is not connected.
Use this feature to display the mA input value (in mA).
Primary Index mA input
All of these parameters are available in RUN mode and used to verify programming. See Readings in RUN Mode on page 23.
The range and values shown for each of these parameters depends on the Operation (P001) chosen. The readings for each operation are listed below.
parameter. These parameters are also available in simulation mode. See the Testing the Configuration section page 85 for instructions on how to control the simulation direction and rate.
Corresponds to the final reading after all programming is applied.
Please note that the following features apply only to the MultiRanger 200: Difference, Average, OCM, and Pump Totalizer.
In general this means that: P920 = (Reading x P061) + P062.
Primary Index Global · Primary Index Level

P001 P050 = 0 P050 ≠ 0
The distance in Units (P005) or % of Span (P007) between Empty (P006) and the monitored surface.
The distance between the monitored surface and Span (P007).
Related • P007 Span
The distance between the monitored surface and the transducer face.

The calculated vessel capacity in Max Volume (P051) or % of Max Volume.
Please note that Volume Measurement is a MultiRanger 200 feature only.
Related • P051 Maximum Volume
The calculated flowrate in Max Flow (P604) units or % of Max Flow.
Please note that Flow Measurement is a MultiRanger 200 feature only.
Related • P604 Maximum Flow
Corresponds to Head (the distance from Zero Head (P605) to the monitored surface in Units (P005) or % of Span (P007).
Please note that Head Measurement is a MultiRanger 200 feature only.

The distance between the surface and the transducer face (displays only as % of Empty).
Use P923 unless the distance information is required in percent.
This feature resets all parameters to original values.
Values See below.
Use this feature prior to initial programming if arbitrary Parameter Values were used during a bench test, or after upgrading the software. Following a Master Reset, complete reprogramming is required.
To perform a Master Reset, access P999 and press CLEAR . C.ALL displays until the reset is complete.
In dual point units, both points can be reset at the same time by setting the index to 00
and pressing the CLEAR keys.
Primary Index Transducer · Primary Index · Global Transducer


Transmit Pulse
The transmit pulse consists of one or more electrical “shot” pulses, which are supplied to the transducer connected to the MultiRanger terminals. The transducer fires an acoustic “shot” for each electrical pulse supplied. After each shot is fired, sufficient time is provided for echo (shot reflection) reception before the next (if applicable) shot is fired. After all shots of the transmit pulse are fired, the resultant echoes are processed. The transmit pulse shot number, frequency, duration, delay, and associated measurement range are defined by parameters P803 and P840 to P852.
Echo processing consists of echo enhancement, true echo selection, and selected echo verification.
Echo Enhancement is achieved by filtering (P821 and P822) and reforming (P823) the echo profile. The true echo (echo reflected by the intended target) is selected when that portion of the echo profile meets the evaluation criteria of Sonic Intelligence®. Insignificant portions of the echo profile outside of the measurement range (Span P006 + Range Extension P801), below the TVT Curve (P830, and P832 to P835), and less than the Confidence Threshold (P804) and Short Shot Floor (P851) are automatically disregarded. The remaining portions of the Echo Profile are evaluated using the Algorithm (P820) and Short Shot Bias (P850). The Echo Profile portion providing the best Echo Confidence (P805) is selected.
True echo verification is automatic. The position (relation in time after transmit) of the new echo is compared to that of the previously accepted echo. When the new echo is within the Echo Lock Window (P713), it is accepted and displays, outputs, and relays are updated per the Fuzz Filter (P710) and Rate Parameters (P700 to P703). If the new echo is outside of the Window, it is not accepted until Echo Lock (P711) requirements are satisfied.


A TVT curve describes a threshold below which any echoes will be ignored. The default TVT curve is used, until P837 and P838 are used to create a new ’learned TVT curve’.
False echoes can be caused by an obstruction in the beam path (pipes, ladders, chains, and such). Such false echoes may rise above the default TVT curve.
P838 allows you to set a distance, and P837 then instructs the MultiRanger to ’learn’ where the obstructions/false echoes are within that distance. The new TVT curve is set above the false echoes, screening them out.
(or when P837 = 0)
Display after Auto False Echo Suppression
TVT curve (learned) material level

To calculate the transducer to material level (object) distance, the transmission medium (atmosphere) sound velocity (P653) is multiplied by the acoustic transmission to reception time period. This result is divided by 2 to calculate the one way distance.
Distance = Sound Velocity x Time / 2
The Reading displayed is the result of performing any additional modification to the calculated distance (as determined by Operation P001, Units P005, Volume Conversion, P050 to P054, Reading, P060 to P063, OCM, P600 to P611, and/or Totalizer P622 to P633 parameters).
The sound velocity of the transmission medium is affected by the type, temperature, and vapor pressure of the gas or vapor present. As preset, the MultiRanger assumes the vessel atmosphere is air at 20°C (68°F). Unless altered, the sound velocity used for the distance calculation is 344.1 m / s (1129 ft / s).
Variable air temperature is automatically compensated when a Siemens Milltronics ultrasonic / temperature transducer is used. If the transducer is exposed to direct sunlight, use a sunshield or a separate TS-3 temperature sensor.
Also, if the temperature varies between the transducer face and the liquid monitored, use a TS-3 temperature sensor (submerged in the liquid) in combination with an ultrasonic / temperature transducer. Set Temp Source (P660) for both, to average the transducer and TS-3 measurements.
Atmosphere composition other than air can pose a challenge for ultrasonic level measurement. However, excellent results may be obtained by performing a Sound Velocity Calibration (P651) if the atmosphere is homogeneous (well mixed), at a fixed temperature, and under consistent vapor pressure.
The MultiRanger automatic temperature compensation is based on the sound velocity / temperature characteristics of “air” and may not be suitable for the atmosphere present. If the atmosphere temperature is variable, perform frequent Sound Velocity Calibrations to optimize measurement accuracy.
Sound Velocity calibration frequency may be determined with experience. If the sound velocity in two or more vessels is always similar, future calibrations may be performed on one vessel and the resultant Velocity (P653) entered directly for the other vessel(s).
If the sound velocity of a vessel atmosphere is found to be repeatable at specific temperatures, a chart or curve may be developed. Then, rather than performing a Sound Velocity Calibration each time the vessel temperature changes significantly, the anticipated Velocity (P653) may be entered directly.
![Page 237 of the Siemens 200 user manual — Volume Calculation [MR 200]](/cdn/reader/82/829c15aaebfbf17c854075ac1e0dbc01f9b82abbfd8c8576c5c8e04c8c601513/images/page-237.webp?w=400)
When echo processing is complete (if more than 1 vessel is monitored) the scanning relay changes state to supply the transmit pulse to the other transducer after the Scan Delay (P727).
Scan Delay is automatically set by Maximum Process Speed (P003). When high speed scanning is required (sometimes the case for equipment position monitoring), the Scan Delay may be reduced. Reduce the Scan Delay only as required, otherwise premature scanning relay fatigue could occur.
When two transducers are connected and configured in a dual point unit, the MultiRanger will scan each in turn via the scanner relay. When a single point MultiRanger is programmed for differential or average level Operation (P001 = 4 or 5), two transducers of the same type must be used.
Please note that the Volume Calculation feature is only available on the MultiRanger 200.
The unit provides a variety of volume calculation features (P050 to P055).
If the vessel does not match any of the eight preset Tank Shape calculations, a Universal Volume calculation may be used. Use the level/volume graph or chart provided by the vessel fabricator (or create one based on the vessel dimensions). Based on the graph, choose the Universal Volume calculation, and select the level vs. volume breakpoints to be entered (32 max). Generally, the more breakpoints entered, the greater the accuracy.
Universal, Linear (P050 = 9)
This volume calculation creates a piece-wise linear approximation of the level/volume curve. This option provides best results if the curve has sharp angles joining relatively linear sections.
Enter a Level Breakpoint at each point where the level/volume curve bends sharply (2 minimum).
For combination curves (mostly linear but include one or more arcs), enter numerous breakpoints along the arc, for best volume calculation accuracy.

Please note that the Universal, Curved feature is only available on the MultiRanger 200.
Set P050 = 10
This calculation creates a cubic spline approximation of the level/volume curve, providing best results if the curve is non-linear, and there are no sharp angles.
![Siemens 200 — Universal, Curved [MR 200] diagram (page 238)](/cdn/reader/82/829c15aaebfbf17c854075ac1e0dbc01f9b82abbfd8c8576c5c8e04c8c601513/images/fig-238-1.webp)
Please note that the Flow Calculation feature is only available on the MultiRanger 200.
The MultiRanger provides numerous OCM flow calculation features (P600 to P611).
If the PMD (primary measuring device) does not match any of the eight preset PMD calculations, or if a PMD is not used, select a Universal Volume calculation. Use the head/ flow graph or chart provided by the PMD fabricator (or create one based on the PMD or channel dimensions).
Based on the graph, choose the Universal Flow calculation, and select the head versus flow breakpoints to be entered (32 max). Generally, the more breakpoints entered, the greater the flow calculation accuracy.

Please note that the Universal, Linear feature is only available on the MultiRanger 200.
Set P600 = 4.
This flow calculation creates a piece-wise linear approximation of the head/flow curve. This option provides best results if the curve has sharp angles joining relatively linear sections.
Enter a Head Breakpoint at each point where the head/flow curve bends sharply (2 minimum). For combination curves (mostly linear but include 1 or more arcs), enter numerous breakpoints along the arc, for best flow calculation accuracy.
See also Typical Flow Characterization on page 83.
Please note that the Universal, Curved feature is only available on the MultiRanger 200.
This calculation creates a cubic spline approximation of the head/flow curve, providing best results if the curve is non-linear, and there are no sharp angles.

For combination curves, enter at least 2 breakpoints immediately before and after any sharp angle (as well as 1 breakpoint exactly at the angle) on the curve. For more information, go to Typical Flow Characterization on page 83.
The MultiRanger’s ability to respond to material level changes is designed to exceed even the most demanding installation requirements.
The Maximum Process Speed setting automatically presets various parameters affecting the MultiRanger response to material level changes as follows:
Speed (P003) (units) 1 (slow) 2 (medium) 3 (fast) P070 Failsafe Timer (min) 100 10 1 P700 Max Fill Rate (m/min) 0.1 1 10 P701 Max Empty Rate (m/min) 0.1 1 10 P702 Filling Indicator (m/min) 0.01 0.1 1 P703 Emptying Indicator (m/min) 0.01 0.1 1 P704 Rate Filter (option) 4 2 2 P710 Fuzz Filter (% of Span) 100 50 10 P713 Echo Lock Window (per P701 / P702 and time since last valid measurement) P727 Scan Delay (seconds) 5 5 3 P841 Long Shot Number 10 5 2
If any of these parameters are independently altered, a Maximum Process Speed (P003) parameter alteration automatically resets the independently altered value.
Slower Maximum Process Speed (P003) provides greater measurement reliability. Faster, independently set Max Fill (P700) and Max Empty (P701). Rates may be impeded by Echo Lock (P711), Scan Delay (P727), and Shot Delay (P728) values.

If the setup becomes too confusing use P999 to reset and start again. Common Problems Chart
Symptom Cause Action Display blank, transducer not pulsing. No power. Check power supply, wiring, or power fuse.
Check programmer usage: 15 cm (6”) from faceplate pointed at upper target.
No response to program- mer.
Obstructed infrared interface, defective programmer.
Short circuited transducer cable, or defective transducer at indi- cated terminal block number.
Displays Short and tb:(#).
Repair or replace as necessary.
Transducer not connected or con- nection reversed.
Check connection to displayed ter- minal blocks. Open circuited transducer cable, or defective transducer at indi- cated terminal block number.
Displays Open and tb:(#).
Repair or replace as necessary.
Relocate and/or re-aim transducer at material. Proceed to Measurement Difficul- ties.
Displays LOE. Weak or non-existent echo.
Wrong transducer selected (P004).
Verify transducer type and re-enter value. Transducer connected in “two wire” method.
Do not tie white and shield together. Use all three terminal blocks. Transducer connected back- wards.
Displays Error and tb:(#).
Reverse black and white wires on terminal block.
Displays EEEE Value too large to display in 4 or 5 characters.
Select larger Units (P005), or lower Convert Reading (P061). Reading fluctuates while material level is still (or vice versa).
Alter Maximum Process Speed (P003) or damping (P704) accord- ingly. See Maximum Process Speed.
Incorrect measurement stabiliza- tion.
Relocate and / or re-aim transducer at material level or object. Proceed to Measurement Difficul- ties below. See also: Transducer Ringing. Material level reported is always incorrect by the same amount.
Transducer acoustic beam obstructed, standpipe too nar- row, or transducer ringing (reads over 100%).
Reading is fixed, regard- less of the actual mate- rial level.
Incorrect Empty (zero) reference for level operation (P001 = 1).
See Empty (P006), Reading Offset (P063), Offset Calibration (P650), & Offset Correction (P652).
Measurement accuracy improves as level nears transducer.
Use a transducer with a built-in tem- perature sensor or a TS-3 tempera- ture sensor. See Sound Velocity.
Incorrect Sound Velocity used for distance calculation.
Reading is erratic, with little or no relation to material level.
Relocate and / or re-aim transducer at material. Check noise parameters. See Noise Problems.
True echo too weak or wrong echo being processed.

Incorrect readings can be the result of noise problems, either acoustic or electrical, in the application.
The noise present at the input to the ultrasonic receiver can be determined by viewing parameter P807. The display reads ##:##, where the first number is the average noise and the second is the peak noise. In general, the most useful value is the average noise.
With no transducer attached the noise is under 5 dB. This is often called the noise floor. If the value with a transducer attached is greater than 5 dB, then signal processing problems can occur. High noise decreases the maximum distance that can be measured. The exact relationship between noise and maximum distance is dependent on the transducer type and the material being measured. Any average noise level greater than 20 dB is probably cause for concern unless the distance is much shorter than the maximum for the transducer.
Disconnect the transducer from the MultiRanger. If the measured noise is below 5 dB, then continue here. If the measured noise is above 5 dB go to Non-Transducer Noise Sources below.
average noise. 3. Remove the positive wire of the transducer. Record the average noise. 4. Re-connect the positive wire and remove the negative wire. Record the average
noise. Using the table below, determine the appropriate next step. The terms higher, lower and unchanged refer to the noise recorded in the previous steps.
These are guidelines only. If the suggested solution does not solve the problem, try the other options also.
higher Reducing Electrical Noise unchanged Common Wiring Problems lower Reducing Acoustical Noise
higher Reducing Electrical Noise unchanged Contact Siemens Milltronics lower Reducing Acoustical Noise
higher Common Wiring Problems unchanged Common Wiring Problems lower Reducing Acoustical Noise
To confirm that the problem is acoustical, place several layers of cardboard over the face of the transducer. If the noise is reduced, the noise is definitely acoustical.

Remove all input and output cables from the MultiRanger individually while monitoring the noise. If removing a cable reduces the noise, that cable may be picking up noise from adjacent electrical equipment. Check that low voltage cables are not being run adjacent to high voltage cables or near to electrical noise generators such as variable speed drives.
Filtering cables is an option but is not recommended unless all other options have been exhausted.
The MultiRanger is designed to work near heavy industrial equipment such as variable speed drives. Even so, it should not be located near high voltage wires or switch gear.
Try moving the electronics to a different location. Often moving the electronics a few meters farther from the source of noise will fix the problem. Shielding the electronics is also an option, but it should be a last resort. Proper shielding is expensive and is difficult to install properly–the shielding box must enclose the MultiRanger electronics completely, and all wires must be brought to the box through grounded metal conduit.
Extension wire must be shielded twisted pair. Older MultiRanger products may have included recommendations to use co-axial cable for noise reduction purposes. This is no longer the case. Use shielded twisted pair. See the installation section for specifications.

If the Failsafe Timer (P070) expires due to a measurement difficulty, LOE flashes alternately with the last known Reading. In rare cases, the MultiRanger may lock on to a false echo and report a fixed or wrong reading.
The loss of echo (LOE) display appears when the echo confidence is below the threshold value set in P805 Echo Confidence.
See the transducer manual for range, mounting, and aiming details. For optimum performance, adjust transducer aiming to provide the best Echo Confidence (P805) and Echo Strength (P806) for all material levels within the measurement range.
The most efficient method of checking echoes is with Siemens Milltronics Dolphin Plus software.
Editing Parameters

Use Dolphin Plus to graphically display the echo profile at the installation. Interpret the echo profile and change relevant parameters.
Edit the parameter values. Use F1 to get online help at any time.

Press and hold for four seconds (Failsafe Time Left changes to the Short:Long Confidence display).
To display Echo Confidence in the program mode, access the Echo Confidence (P805) parameter.
To update the value displayed after each aiming adjustment..
Press (five times or more to verify stability and overcome any echo lock P711).
Increase the Failsafe Timer (P070) value, if failsafe operation will not be compromised by the larger value.
Try this only if LOE shows for short periods of time.
Sometimes the interference echoes from the sides of a vessel can cause the MultiRanger to lock onto a consistent, incorrect level. Try installing a longer range (narrower beam) transducer, enter the new transducer model (P004), and (if necessary) optimize aiming and frequency again.
Always contact your Siemens Milltronics service personnel before selecting a transducer to solve this type of problem.
If a narrower beam transducer is not available, use Dolphin Plus to view live sonic profiles and make adjustments to the Advanced Echo Processing parameters. Fixed Reading
If the Reading is a fixed value, regardless of the transducer to material surface distance, ensure the:
stopped, ensure that the mixer blade is not stopped under the transducer.
Check for (and remove if present) any acoustic beam obstruction, or relocate the transducer.
If an obstruction cannot be removed or avoided, adjust the Time Varying Threshold (TVT) Curve to reduce the Echo Confidence derived from the sound reflected by the obstruction. Use Dolphin Plus to adjust the TVT curve. (See P832: TVT Shaper.) Nozzle Mountings
If the transducer is mounted on or in a nozzle, grind smooth any burrs or welds on the inside or open end (the end that opens into the vessel). If the problem persists, install a

larger diameter or shorter length nozzle, bevel the inside of the bottom end, or cut the open end of the nozzle at a 45° angle.
See the transducer manual for complete mounting instructions.
For ST-series and XPS-10 transducers use the plastic conduit / flange adapter supplied with the unit.
If the mounting hardware is over tightened, loosen it. Over tightening changes the resonance characteristics of the transducer and can cause problems. Set the MultiRanger to Ignore the Bad Echo
If the preceding remedies have not fixed the problem, the false echo has to be ignored.
If the Echo is Close to the Transducer If there is a static, incorrect, high level reading from the MultiRanger there is probably something reflecting a strong echo back to the transducer. If the material level never reaches that point extend the Near Blanking (P800) to a distance to just past the obstruction.
Adjust the TVT to Ignore the Echo Use Auto False Echo Suppression (P837-P839) to automatically shape around false echoes. Wrong Reading
If the Reading is erratic, or jumps to some incorrect value periodically, ensure the:
If a periodic wrong Reading is always the same value, see Fixed Reading.
If the wrong Reading is random, ensure the material surface to transducer distance is less than the Empty value entered plus 20%. If the material/object monitored is outside this distance, increase Range Extension (P801) as required. This error is most common in OCM applications using weirs. Liquid Splashing
If the material monitored is a liquid, check for splashing in the vessel. Enter a lower Maximum Process Speed (P003) value to stabilize the Reading, or install a stilling well. (Contact Siemens Milltronics or your local distributor.) Adjust the Echo Algorithm
Use Dolphin Plus to view live sonic profiles and make adjustments to the P820 Algorithm parameter. See P820 on page 204 for details.
If the “Area” algorithm is used and narrow noise spikes are evident on the (long shot) Echo Profile, turn the Spike Filter (P821) on and/or widen the Narrow Echo Filter (P822). Also, if the true echo has jagged peaks, use Reform Echo (P823).

If multiple echoes appear on the Echo Profile, typical of a flat material profile (especially if the vessel top is domed), use the “first” Algorithm.
If the Echo Profile repeatedly switches from short to long, adjust the Short Shot Range (P852) to stabilize the “shot” mode used for the echo evaluation. Also, adjust the Short Shot Bias to increase (or decrease the amount of preference given to short shot echoes over long shot echoes.
Should a stable measurement still not be attainable, contact Siemens Milltronics or your local distributor. Transducer Ringing
If the transducer is mounted too tightly, or if it is mounted so that its side touches something, its resonance characteristics change and this can cause problems.


Normal Ring Down

Ring down times that extend past the near blanking area can be interpreted by the MultiRanger as the material level and are characterized by a steady high level being reported.
All changes and repairs must be done by qualified personnel and applicable safety regulations must be followed. Please note the following:

Please note that some pump control features apply to the MultiRanger 200 only.
The MultiRanger has the pump control strategies to solve nearly any water / wastewater application. This section details these strategies for engineers requiring in-depth knowledge of the system and how it operates.
The various methods of pump control are made up of a combination of two control variables:
The pump duty indicates in what sequence pumps are started.
The start method indicates whether new pumps start and run with any currently running pumps (most common) or whether new pumps start and shut off currently running pumps.
The MultiRanger groups pumps that use identical pumping strategies based on the value of Relay Control Function (P111). Generally, one group of pumps corresponds to one wet well or reservoir.
Please note that this feature applies to the MultiRanger 200 only.
To trigger pump starts by the rate of change in material level use P121–Pump by Rate (P121). New pumps are started, one at a time, until the rate setpoint (Filling Indicator (P702), or Emptying Indicator(P703)) is reached.
Please note that the MultiRanger 100 and the MultiRanger 200 use this feature differently.
All of these algorithms can be used to start multiple pumps (assist).

All of these algorithms can be used to start multiple pumps (assist) or one pump at a time (backup).
The MultiRanger 100 and MultiRanger 200 have three main methods of pump control:
Starts pumps based on individual setpoints and always starts the same pumps in the same sequence.
Starts pumps based on the duty schedule and always leads with a new pump.
Starts pumps based on user-defined ratio of running time.
Ties the indexed pump relay directly to the indexed setpoint.
The relay contact closes at the ON setpoint and opens at the OFF setpoint. Multiple relay contacts in the pump group can be closed at the same time.
The following table shows relay status when each setpoint is reached.
Index 1 2 3 On 3 On 2 On On Off On 1 On Off Off Off 0 Off Off Off
Fixed Duty Backup (P111 = 51) [MR 200]
Ties the indexed pump relay directly to the indexed setpoint.
Please note that this feature applies to the MultiRanger 200 only.
The relay contact closes at the ON setpoint and opens at the OFF setpoint. When a new relay trips the previously closed relay contact opens to shut down the running pump.
Only one relay contact in the pump group can be closed at any one time.

The following table shows relay status when each setpoint is reached.
Index 1 2 3 On 3 Off Off On On 2 Off On Off On 1 On Off Off Off 0 Off Off Off
Alternate Duty Assist (P111 = 52)
Alternates the lead pump each time the material level cycles and runs all pumps together.
The setpoints associated with the relays are grouped so that they can be rotated.
Setpoint one does not relate directly to relay one. The pumping algorithm manages the mapping of setpoints to relays.
When pumps are run, they RUN in parallel.
Cycle 1 Relays
Setpoints
On 3 On 2 On On Off On 1 On Off Off Off 0 Off Off Off Cycle 2 Relays
On 3 On 2 Off On On On 1 Off On Off Off 0 Off Off Off Cycle 3 Relays
On 3 On 2 On Off On On 1 Off Off On Off 0 Off Off Off
Alternates the lead pump each time the material level cycles.
Please note that this feature applies to the MultiRanger 200 only.
The setpoints associated with the relays are grouped so that they can be rotated. Setpoint one does not relate directly to relay one. The pumping algorithm manages the mapping of setpoints to relays. When pumps are run, they can RUN only one at a time.
![Page 252 of the Siemens 200 user manual — Relay Operation (for P118 = 2) [MR 200]](/cdn/reader/82/829c15aaebfbf17c854075ac1e0dbc01f9b82abbfd8c8576c5c8e04c8c601513/images/page-252.webp?w=400)
Cycle 1 Relays 1 2 3
On 3 Off Off On On 2 Off On Off On 1 On Off Off 0ff 0 Off Off Off
Cycle 2 Relays 1 2 3
On 3 On Off Off On 2 Off Off On On 1 Off On Off Off 0 Off Off Off Cycle 3 Relays 1 2 3
On 3 Off On Off On 2 On Off Off On 1 Off Off On Off 0 Off Off Off
Selects the lead pump based on number of hours each pump has run and the specified ratios that each pump requires. Multiple pumps can run at one time.
Please note that this feature applies to the MultiRanger 200 only.
Please note that this feature applies to the MultiRanger 200 only.
The setpoints associated with the relays are grouped so they can be redistributed based on pump RUN time ratios. The next pump to start or stop is the one with the required time to actual time ratio.
Over time the number of hours demanded of each pump will conform to the ratios specified. Usually, the ratios are specified in percent values.
To create a grouping of pumps where two pumps make up 50% of the run time and the third pump makes up the other 50%.
P122 is set to these values:
P122 Index Value 1 25 2 25 3 50

Selects the lead pump based on number of hours each pump has RUN and the specified ratios that each pump requires. Only one pump can RUN at a time.
Please note that this feature applies to the MultiRanger 200 only.
This algorithm is the same as Service Ratio Duty Assist except that it will only RUN one pump at a time. When the next pump in the sequence starts, the previous pump stops.
Selects the lead pump based on the Alternate duty but uses staggered off setpoints and shuts down pumps based on the first in, first out rule.
Please note that this feature applies to the MultiRanger 200 only.
This algorithm starts pumps in the same way as Alternate Duty Assist but uses staggered OFF setpoints to shut the pumps down. When the first OFF setpoint is reached the FIFO rule shuts down the first pump started. If the pumps started in sequence 2,3,1 then they would be shut down in sequence 2,3,1.
Starts pumps until the level is changing at the rate specified in P702 or P703.
Please note that this feature applies to the MultiRanger 200 only.
Pumping costs can be less because only the highest ON setpoint needs to be programmed and this results in a lower difference in head to the next wet well which, in turn, results in less energy being used to pump out the well.
Please note that these features apply to the MultiRanger 200 only.
There are a number of other controls available to modify pump behaviour.
Extends the RUN period for a pump based on a set time interval. This allows for the wet well to be pumped lower than usual and reduces sludge build-up on the well bottom.
Varies the ON and OFF setpoints to keep a fat ring from forming around the walls of the wet well.
Allows for two different Alternate Duty Assist or Alternate Duty Backup pump groups in the same application.

Operates a flush valve or special flush device based on the number of pump starts, usually to aerate wet well wastewater.

Updating Software
To update the software in the MultiRanger, please contact your Siemens Milltronics representative to obtain the software. For a complete list of representatives, go to www.siemens.com/processautomation
To install the software, please complete the following:
settings to match your MultiRanger. Please note that the software default settings will already match the default settings of the unit. Changes are only necessary if the RS-232 settings in the unit have been changed. 3. Complete the Downloader program steps. 4. Verify that Downloader confirms a succesful upgrade before exiting. 5. Complete a Master Reset (P999) after a successful upgrade, before re-entering

The following procedure will assist you if you are upgrading from a MultiRanger Plus to a MultiRanger 100 or MultiRanger 200.
If the application is unchanged, copy down the parameters in the MultiRanger Plus before de-commissioning the unit.
Please read the MultiRanger Installation section on page 8 before mounting the new MultiRanger 100/200.
Make sure the power is OFF to the unit before following these steps:
If this is not practical, please refer to the Coaxial Transducer Extension section below.
The MultiRanger 100/200 uses a new differential input receiver that works either directly connected to the transducer lead or with a screened twisted pair extension cable via a field junction box. This arrangement, using two conductors and a screen, gives considerably better electrical noise immunity than the previous coaxial arrangement (up to 20dB) and will give reliable operation in applications where the proximity of power cables, variable speed drives etc. would have caused problems.
If you are installing a new MultiRanger system we strongly recommend that you use a good quality screened, twisted pair cable if an extension is required. If the integral transducer cable is used you need only connect to the MultiRanger and benefit from the superior performance.
Transducer

If you are replacing an older Siemens Milltronics Ultrasonic Level Controller with a new MultiRanger where an RG62 coaxial extension is fitted and you are unable to replace the extension with a new cable, please refer to connection diagram below. Please note that the noise immunity performance will be similar to our older model ultrasonic level controllers if you use coaxial cable.
RG62 coaxial cable has been used to make the extension. The level controller might be a MultiRanger Plus, HydroRanger, HydroRanger Plus or one of our other level controllers.
Connect with a 0.1µF (100v or greater) capacitor (included with MultiRanger) between the shield and negative terminals. Connect centre core of the coaxial to the positive terminal and the screen to the negative terminal.
Transducer · Transducer

The MultiRanger 100/200 uses the current Siemens Milltronics standard parameter set which is different than the parameter numbers used in the MultiRanger Plus.
The MultiRanger Plus parameters numbers are sequential (P-0 to P-99). The MultiRanger 100/200 parameters number are also sequential; however, some of the parameters are indexed.
Parameters are indexed when they apply to more than one input or output. The index value defines the input/output for that parameter. Indexed parameters contain a value for each index, even if that index is not used.
In the MultiRanger Plus each relay had a dedicated parameter number for its function, ON and OFF points.
MultiRanger Plus Relay One
MultiRanger 100/200 uses P111 for all relays and indexes these parameters.
The following tables show the differences between the MultiRanger Plus and the MultiRanger 100/200. They each show one relay programmed as a high alarm, and two relays programmed for pump control with alternating pump starts:
MultiRanger 100/200
Relay Parameter Value
Relay Parameter Value
1 P-8 (relay function) 1 (alarm)
1 P111 [1] (relay function) 1
1 P-9 (ON setpoint) 4 m
1 P112 [1] (ON setpoint) 4 m
1 P10 (OFF setpoint) 3.5 m
1 P113[1] (OFF setpoint) 3.5 m
2 P-11 (relay function) 9
2 P111[2] (relay function) 52
2 P-12 (ON setpoint) 3.2 m
2 P112[2] (ON setpoint) 3.2 m
2 P-13 (OFF setpoint) 0.5 m
2 P113[2] (OFF setpoint) 0.5 m
3 P-14 (relay function) 9
3 P111[3] (relay function) 52
3 P-15 (ON setpoint) 3.4 m
3 P112[3] (ON setpoint) 3.4 m
3 P-16 (OFF setpoint) 0.6 m
3 P113[3] (OFF setpoint) 0.6 m


7ML19981FBO6 MultiRanger 100/200 – INSTRUCTION MANUAL Page 249


Page 250 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FBO6

MultiRanger 100
Date Changes
Rev.
1.00 Aug. 15, 2005 • Pilot Release.
1.01 Sept. 21, 2005 • Internal Release.
1.02 Dec. 12, 2001 • Initial Release.
Appendix H: Software Rev.

Date Changes
Rev.
Feb. 11, 2003
Feb. 23, 2003
1.06 May 27, 2004 • RCVR offset limits are no longer handled in the code.
Appendix H: Software Rev.

Date Changes
Rev.
Appendix H: Software Rev.
1.10 not released

Date Changes
Rev.
1.00 Nov. 09, 2001 • Pilot Release.
1.01 Dec. 02, 2001 • Initial Release.
Appendix H: Software Rev.
1.03 Sept. 25, 2002 • Allow files to be downloaded to the product using Dolphin+.

Date Changes
Rev.
Feb. 11, 2003
Feb. 21, 2003
1.06 May 27, 2004 • RCVR offset limits are no longer handled in the code.
Appendix H: Software Rev.

Date Changes
Rev.
Appendix H: Software Rev.
1.10 not released
1.12 Aug. 01, 2006 • Internal Release.

# Name 1 2 3
P003 Max. Process Speed
P052 Tank Dimension A
P053 Tank Dimension L
P054 Level Breakpoints record values on a separate sheet.
P055 Volume Breakpoints record values on a separate sheet.
P064 Reading Override Enable
P065 Reading Override Value
P066 Override Time Delay
P071 Failsafe Material Level
P072 Failsafe Level Advance
P100 Preset Applications [MR200]
P111 Relay Control Function
P112 Relay ON Setpoint

# Name 1 2 3
P113 Relay OFF Setpoint
P115 Relay Interval Setpoint [MR 200]
P118 Relay Output Logic
P119 Relay Logic Test
P121 Pump by Rate [MR 200]
P122 Pump Service Ratio [MR 200]
Advanced Pump Control Modifiers
P130 Pump Run-On Interval [MR 200]
P131 Pump Run-On Duration [MR 200]
P132 Pump Start Delay [MR 200]
P133 Pump Power Resumption Delay
P136 Wall Cling Reduction [MR 200]
P137 Pump Group [MR 200]
P170 Flush Pump [MR 200]
P171 Flush Cycles [MR 200]
P172 Flush Interval [MR 200]
P173 Flush Duration [MR 200]
P200 mA Output Range
P201 mA Output Function
P202 mA Output Allocation
P203 mA Output Value / Transducer
P210 0/4 mA Output Setpoint
P211 20 mA Output Setpoint
P212 mA Output Min Limit
P213 mA Output Max limit
P214 4 mA Output Trim
P215 20 mA Output Trim

# Name 1 2 3
P219 mA Output Failsafe[MR 200]
P250 mA Input Range [MR 200]
P251 0 to 4 mA Input Level [MR 200]
P252 20 mA Input Level [MR 200]
P253 Input Filter Time Constant [MR 200]
P254 Scaled mA Input Value[MR 200]
P260 mA Raw Input [MR 200]
P270 Discrete Input Function
P275 Scaled Discrete Input Value
P300 Temp, Transducer max
P301 Temp, Transducer min
P302 Temperature, Sensor max
P303 Temperature, Sensor min
P309 Pump RUN Time
P312 Pump Run Ons [MR 200]
P320 Flow Max [MR 200]
P321 Flow Min [MR 200]
P322 LCD Total Low [MR 200]
P323 LCD Total High [MR 200]
P331 Auto Record Enable
P332 Auto Record Transducer
P333 Auto Record Interval
Auto Record ON and OFF Setpoints

# Name 1 2 3
P334 Auto Record ON Setpoint
P335 Auto Record OFF Setpoint
P336 Auto Record Filling/Emptying
P337 Auto Record LOE Time
P340 Date of Manufacture
P600 Primary Measuring Device [MR 200]
P601 Flow Exponent [MR 200]
P602 Primary Measuring Device Dimensions [MR200]
P603 Maximum Head [MR 200]
P604 Maximum Flow [MR 200]
P605 Zero Head [MR 200]
P606 Time Units [MR 200]
P607 Flowrate Decimal [MR 200]
P608 Flowrate Units [MR 200]
P610 Head Breakpoints [MR 200]
P611 Breakpoint Flowrates [MR 200]
P620 Low Flow Cutoff [MR 200]
P621 Auto Zero Head [MR 200]
P622 Inflow / Discharge Adjust [MR 200]
P630 LCD Totalized Multiplier [MR 200]
P633 LCD Totalized Decimal Postion [MR 200]
P640 Remote Totalized Multiplier [MR 200]
P641 Flow Sampler Mantissa [MR 200]
P642 Flow Sampler Exponent [MR 200]
P645 Relay Duration [MR 200]

# Name 1 2 3
P651 Sound Velocity Calibration
P654 Velocity @ 20° C
P663 Temp Transducer Allocation
P700 Max Fill Rate
P701 Max Empty Rate
P705 Rate Update Time
P706 Rate Update Distance
P708 Volume Rate Display [MR 200]
P712 Echo Lock Sampling
P713 Echo Lock Window
P726 Level System Sync
P731 Auxiliary Reading Key

# Name 1 2 3
P737 Primary Reading [MR 200]
P779 Modem Inactivity Timeout
P782 Parameter Index Location
P791 Hardware Error Code
P792 Hardware Error Count
P794 SmartLinx Module Type
P802 Transducer with Submergence Shield
P803 Shot / Pulse Mode
P815 Echo Time Filtered
P816 Echo Time Raw
P822 Narrow Echo Filter

# Name 1 2 3
P825 Echo Marker Trigger
P832 TVT Shaper Adjust
P833 TVT Start Min
P834 TVT Start Duration
P835 TVT Slope Min
P837 Auto-False Echo Suppression
P838 Auto-False Echo Suppression- Distance
P839 TVT Hover Level
P840 Short Shot Number
P841 Long Shot Number
P842 Short Shot Frequency
P843 Long Shot Frequency
P844 Short Shot Width
P845 Long Shot Width
P850 Short Shot Bias
P851 Short Shot Floor
P852 Short Shot Range
P900 Software Revision Number
P911 mA Output Value
P914 mA Input [MR 200]

# Name 1 2 3
P924 Volume Measurement [MR 200]
P925 Flow Measurement [MR 200]
P926 Head Measurement [MR 200]








Siemens Milltronics Process Instruments Inc. 1954Technology Drive, P.O. Box 4225 Peterborough, ON, Canada K9J 7B1 Tel: (705) 745-2431 Fax: (705) 741-0466 Email: techpubs.smpi@siemens.com
Siemens Milltronics Process Instruments Inc. 2009 Subject to change without prior notice
Printed in Canada
Pages referenced in this manual — manufacturer, support and downloads.
What the fault codes shown on the display mean and how to clear them.
Common problems, likely causes, and what to do.
| Problem | Possible cause | Remedy | |
|---|---|---|---|
| Display blank, transducer not pulsing. | No power. | Check power supply, wiring, or power fuse. | p.241 |
| No response to program- mer. | Obstructed infrared interface, defective programmer. | Check programmer usage: 15 cm (6”) from faceplate pointed at upper target. | p.241 |
| Displays Short and tb:(#). | Short circuited transducer cable, or defective transducer at indi- cated terminal block number. | Repair or replace as necessary. | p.241 |
| Displays Open and tb:(#). | Transducer not connected or con- nection reversed. | Check connection to displayed ter- minal blocks. | p.241 |
| Displays Open and tb:(#). | Open circuited transducer cable, or defective transducer at indi- cated terminal block number. | Repair or replace as necessary. | p.241 |
| Displays LOE. | Weak or non-existent echo. | Relocate and/or re-aim transducer at material. | p.241 |
| Displays LOE. | Proceed to Measurement Difficul- ties. | p.241 | |
| Displays Error and tb:(#). | Wrong transducer selected (P004). | Verify transducer type and re-enter value. | p.241 |
| Displays Error and tb:(#). | Transducer connected in “two wire” method. | Do not tie white and shield together. Use all three terminal blocks. | p.241 |
| Displays Error and tb:(#). | Transducer connected back- wards. | Reverse black and white wires on terminal block. | p.241 |
| Displays EEEE | Value too large to display in 4 or 5 characters. | Select larger Units (P005), or lower Convert Reading (P061). | p.241 |
| Reading fluctuates while material level is still (or vice versa). | Incorrect measurement stabiliza- tion. | Alter Maximum Process Speed (P003) or damping (P704) accord- ingly. See Maximum Process Speed. | p.241 |
| Reading is fixed, regard- less of the actual mate- rial level. | Transducer acoustic beam obstructed, standpipe too nar- row, or transducer ringing (reads over 100%). | Relocate and / or re-aim transducer at material level or object. | p.241 |
| Reading is fixed, regard- less of the actual mate- rial level. | Proceed to Measurement Difficul- ties below. | p.241 | |
| Reading is fixed, regard- less of the actual mate- rial level. | See also: Transducer Ringing. | p.241 | |
| Material level reported is always incorrect by the same amount. | Incorrect Empty (zero) reference for level operation (P001 = 1). | See Empty (P006), Reading Offset (P063), Offset Calibration (P650), & Offset Correction (P652). | p.241 |
| Measurement accuracy improves as level nears transducer. | Incorrect Sound Velocity used for distance calculation. | Use a transducer with a built-in tem- perature sensor or a TS-3 tempera- ture sensor. | p.241 |
| Measurement accuracy improves as level nears transducer. | See Sound Velocity. | p.241 | |
| Reading is erratic, with little or no relation to material level. | True echo too weak or wrong echo being processed. | Relocate and / or re-aim transducer at material. | p.241 |
| Reading is erratic, with little or no relation to material level. | Check noise parameters. See Noise Problems. | p.241 |