15 Discovery Way, Acton, MA 01720 Phone: (978)263-3 584, Fax: (978)263-50 86 Web Site: www.princetoninstruments.com Operating Instructions A cton Se ries SP - 2300i 0.300 Meter Focal Length Triple Grating Imaging Monochromator / Spectrograph Rev 4 .0 9.
1 A cton SP-2300i Operating Inst ructions CONTENTS: Page I. Brief Introduction and SP- 2300i Description 2 II. Specifications 3 III. SP - 2300i Setup 4 A. Unpacking Note 4 B. Connecting the SP -2300i to a Computer 4 C. Cables and Connecti ons for SP-2300i 5 D.
2 I : Brief Introductio n and SP-2300i Descr iption Brief Introduction: This instruction manual is i ntended to assist you in s et - up and operation of your new SP -2300i monochromator/spectrograp h.
3 II : SP -2300i Specifications Every SP-2 300i monochromator or spec trograph includes a triple grating turret for 1, 2, or 3 gratings, 32 -bit microprocessor controll ed scanning, buil t-in RS - 232 and USB interfaces, and micrometer controlled e ntrance slit.
4 III : SP -2300i Setup Section III-A: Unpacking and Inspection Carefully unpack and examine the SP-2300i and any ac cessories purc hased. Note: Report any damage i mmediately t o the carrier and t o Princeton Instrum ents, and save all pac ki ng materia l.
5 Section III-C: Cables and Connections for SP-2300i The following computer cables are supplied wi th the SP-2300i: CC - 499 - 2 IBM AT or compatible 9 pin female ( DB9S) connector to 9 pin male connector (DB9P). 3650 -USB- 06 USB Cable Type A to Type B.
6 Section III-D: Mounting Accessories to the SP -2300i Slit A sse mblies All Acton accessories come with thei r own set of instructions for proper mounting and operation. The instructions below are only general i nformation. Pleas e refer to the individual instructions for detailed information.
7 Section III-E: Detector Mounting Instructions: Mounting Focal Plane Detectors to the SP-2300i Spectrograph The standard mounting f lange for CCDs and diode arrays acc ommodates detectors with two diff erent bolt circles. There are three (3) equal ly spaced #10 - 32 tapped holes on a 3.
8 Section III-F: Slit Width A djustments in the Model 716 Bilateral Slits The slit width of eac h bilateral slit assembly (716 type) is adjustable f rom 0.010 millimeters to 3 millimeters (10 to 3,000 µm) by a micrometer knob loc ated on the top of the sl it housing.
9 IV : SP-2300 i Operation Section IV-A: Initialization W hen power is turned ON to the SP -2300i, it ini tializes to a wavelength of 0.0 nm for grating number 1. If the power is switched OFF and then ON again to the SP-230 0i, it will re-ini tialize.
10 NM/MIN Sets the scan rate i n nm/min to 0.01 nm/min resolution with units nm/min ?NM/MIN Returns present scan rate i n nm/min to 0.01 nm/min resoluti on with units nm/min Grating Contro l Commands : GRAT ING Places specified grating in position to the wavelength of the wavelength on the present grating.
11 Slit Width Cont rol Command s (for Motorized Slits only ): FRONT-EX IT-SLIT Designates front ex it slit to receive slit control commands. Note : The designation remains in effec t until changed by another slit des ignator. This command does not have to be repeated unti l the designate d sl it is changed.
12 Section IV-C : Focusing and A lignment of A rra y Detectors W ith the array detector properly mounted to the SP-2300 i, use the followi ng procedure to align and foc us the array detector to the SP-230 0i optical system. It is as sumed that the array detection s ystem is running with SpectraSense or W inSpec.
13 Section IV-D : Conv erting the SP-2300i from Monochromator to Spectrograph, or Spectrograph to Monochromator Monochromator to Sp ectrograph: This procedure requires that you ha ve the monochromator version, Model SP- 2355 or SP- 2357 , plus the array detector mounting fl an ge.
14 V : Ap pendices and Schemat ic Dra wings A ppendix V -A: Selecting Alternate Start-Up Para meters The following are the Start-U p parameters and their def ault values: GRATING #1 WAVELENGTH 0.0 nm SCAN SPEED 100.0 nm/min Each of the above may be changed through the RS - 232 port or USB port using the f ollowing commands.
15 A ppendix V-B: Cable Connection Diagrams - Monochromators & Spectrographs to Computer Figure 9 . RS -232 Cable Wir ing Figure 10 . Cable Connections (Back Panel ) POWER Power connection +5 and +24 Vol ts USB HUB USB Hub 1.1 compatible. USB USB connection to computer.
16 A ppendix V -C: Standard Slit Assembly Drawing.
17 A ppendix V -D: Accessories Listing SP -2300i M onochromators & Spectrographs SP - 2355 SP - 2300i, 300mm Monoch romator: Micrometer adjustabl e entrance/exit slits, 90° optical path, internal.
18 TDS- 429 Dual Light Source with 30 watt deuterium & 30 watt tungsten -halogen lamps. Inc ludes housing with light c ollection mirror, mounting flange, and power supply. Provides c ombined output from 190 to >2.0 µm. XS - 43 3 75 Watt Xenon Light Source includes power s upply, housing, mounting flange and light collection mirror.
19 Filter Wheels and Order Sorting Filter Ass emblies FA - 2448 Six Position Filter Wheel Assembl y. Manual sample index ing, holds up to six samples, each 1.
20 A ppendix V -E: Gratings for SP-2300i Grooves/mm Blaze Wavelength Standard Part Number Mechanical Limit Notes: † Bl az ed H ol og rap hi c, 19 0- 450 nm , op tim um ra ng e †† H ol ogr ap hi .
21 A ppendix V -F: Certification and Warranty Certification Princeton Instruments certifies that thi s instrument was thoroughly tested and f ound to meet the specif ications furnished by Princ eton Instruments when it was shipped f rom the factory.
22 A ppendix V -G: Drawings.
23.
24.
An important point after buying a device Princeton Digital (USA) SP-2300i (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Princeton Digital (USA) SP-2300i yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Princeton Digital (USA) SP-2300i - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Princeton Digital (USA) SP-2300i you will learn all the available features of the product, as well as information on its operation. The information that you get Princeton Digital (USA) SP-2300i will certainly help you make a decision on the purchase.
If you already are a holder of Princeton Digital (USA) SP-2300i, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Princeton Digital (USA) SP-2300i.
However, one of the most important roles played by the user manual is to help in solving problems with Princeton Digital (USA) SP-2300i. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Princeton Digital (USA) SP-2300i along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center