PIONEER .
1 Read this manual before using con centrator and save for future r eference..
2 T ABLE OF CONTENTS 0. IEC SYMBOLS ………...………………………………………….…………………3 1. CAUTIONS …………………..……………………..……………. ………………..3 2. W AR NINGS ………………….
3 0. IEC SYMBOLS Caution, Attention or Consult accompanying documents. Alternating Current T y pe BF Equipment No Smoking or Naked Flam es Use No Oil or G r ease 1. CAUTIONS Federal Law restricts this device fo r sale or use on the ord er of a physician.
4 2. W ARNINGS Fire Hazard-Oxy g en promotes rapid burning. No Smoking or O pen Flames. Electrical Shock Hazard - do not remove the ca binet or open the cover. Disconnect the power cord fr om the electrical outlet before c leaning or ser vicing.
5 3. PRECAUTIONS Alway s plug in a 15 AMP, gr ounded circuit. DO NOT USE AN EXTENSI ON CORD Place the unit where sides and bottom are unobstructed to ensure pr oper airflow. (B e sure that the machine is placed away from drapes or other objects that might block the flow of incoming air and/or alarm indicators.
6 5. DESCRIPTION .
7 6. SPECIFICA TIONS Electrical Requir eme nts 1 15V AC±10 , 60Hz / 230V AC± 10 , 50Hz Rated Curr ent Input 3.5A (1 15V AC/60Hz); 2.0A (230V AC/50Hz) Sound L evel 45dB(A) A verage Operating Altitude Do not use at altitude over 4000 meters. Oxygen Concentration 90 at 1~5L PM * S tated performance maxim um achieved after 60 min.
8 Cabinet I mpact Resistant flame-retarda nt plastic cabinet Regulatory Listing CSA 601 / UL 2601/ IEC/EN-60601-1 Change in Max. Flow with 7 kpa Back Pre ssure 0.3 LPM Recom mended S torage and Shippi ng T emperatur e Dry (Up to 95% non-c ondensing), well ventilated are a -20 ~ 60 (1).
9 7. HANDLING The concentrator should alwa y s be kept in the upright position to prevent cabinet da mage while being transported. The shipping container has been de signed to assure maximum protection of the concentrator . I f the concentrator is to be re shipped by common carrie r , additional cartons are available from Merits.
10 8. OPERA TION 8.1 SELECT A LOCA TION Select a location where sides and bottom are unobstructed to ensure proper air flow . Be sure that the machine is placed away from drapes or other objects that might block the flow of incoming air and/or alarm indicators.
11 / - 7 2 ; FI GURE 2 HUMI DI F.
12 8.5 TURN ON POW ER SWITCH Press the p ower switch to the I position. The p ower indicator light will illuminate. 8.6 ADJU ST FLOWRA TE T urn the knob on the flowmeter to adjust the flowrate to that prescribed by y our physician. The ball inside the flowmeter must be centered on the f lowline for the desired flowrate.
13 9. ALARMS The concentrator is equipped with a visual and a n audible indicator . POWER LOSS: An audible alarm will sound and the power indicator light will turn of f.
14 4. The oxy gen conc entration should be checked eve ry 90 day s for units without optional oxyg en sensors. The oxy gen conc entration should be checked eve ry 6 months for units with optional oxyg en sensors. 10.2 CLEANING THE CABINET FIL TERS NOTE: All conce ntrator models have one cabinet filter located at r ear side of t he cabinet.
15 10.4 HUMIDIFIER NOTE: T o clean the oxyg en humidif ier , follow the instructions prov ided by t he manufactur er . If none ar e pr ovided, foll ow these steps 1. Clean the humidifier every day . 2. W ash it in soapy water and rinse it with a solution of 10 parts water and 1 part vineg ar .
16 W a rranty Merits Health Products Co., Ltd. war rants the Pioneer Ox y gen Conce ntrator to be free f rom defective workmanship and materials for a per iod of 5 y ears fr om the date of purchase. Chec k with y our dealer f or more details. This warranty is limited to the original purchaser of the O xyg en Concentrator .
An important point after buying a device Pioneer Q150 (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Pioneer Q150 yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Pioneer Q150 - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Pioneer Q150 you will learn all the available features of the product, as well as information on its operation. The information that you get Pioneer Q150 will certainly help you make a decision on the purchase.
If you already are a holder of Pioneer Q150, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Pioneer Q150.
However, one of the most important roles played by the user manual is to help in solving problems with Pioneer Q150. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Pioneer Q150 along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center