Philips Patient Monitor (M2540)

220 pages·4.4 MB·Audio System·PDF·ENDEFRITNLJA
Philips Patient Monitor (M2540)
Table of Contents 287 sections 1 About This Manual · About This Manual · Audience · Format · … ShowHide
1 About This Manualp.1About This Manualp.1Audiencep.2Formatp.3Conventionsp.3Service Manual Questions or Commentsp.3Customer Assistancep.41 General Informationp.6Product Overviewp.6System Descriptionp.7Cartp.7Cart Wheelsp.7PCp.7System Monitorp.8e-boxp.8System Control Panelp.8Physio Modulep.9Power Supplyp.9Optional Peripherals (VCR, Printers, and Image Devices)p.9Preset Functionsp.10Keyboard Equivalenciesp.11Safety Specificationsp.12Safety Limitsp.12Regulatory Compliancep.122 Specificationsp.13Physical Dimensionsp.14Electrical Specificationsp.19Monitorp.19Transducer Specificationsp.20Connection and Communication Specificationsp.21System I/O Portsp.21PC Portsp.21I/O Panel Portsp.23Physio Port Specificationsp.23Audio/Video Specificationsp.23Audiop.24External Videop.24Video Output Specificationsp.253 Safetyp.26Transmit Power (Acoustical)p.27Acoustic Exposurep.27AIUM/NEMA Output Display Standardp.28Automatic Index Selectionp.29When MI is Displayed With the Normal Settingp.29When TIS is Displayed With Normal Settingp.29When TIC is Displayed With Normal Settingp.30Explosive Hazardsp.30Electrical Warningsp.31Peripheral Connectionsp.31Glutaraldehyde Exposurep.32Moving the Systemp.32Before Moving the Systemp.32When Moving the Systemp.33Moving on Ramps or Uneven Surfacesp.33System Tiltingp.33Transporting the System in a Vehiclep.34Electromagnetic Compatibilityp.35Avoiding Electromagnetic Interferencep.35Restrictions for Usep.36Immunity Level Test Resultsp.36Electrostatic Dischargep.36Electrosurgical Unitsp.374 Theory of Operationp.38Overviewp.38Internal PCp.39Standard PC Componentsp.39PC Portsp.39VCR-Option Components in the PCp.40Optional Drive-Bay Componentsp.40TR Boardsp.40BPAP Boardp.42Omni TEE Circuitryp.42Demodulator Boardp.43APIO Boardp.43Physio Module Optionp.43Primary Functionsp.44Connectorsp.44System Motherboardp.44Distribution Board and Connector Modulesp.44Transducer Connector Module Typesp.45I/O Panelp.45System Control Panelp.45Power Supplyp.46Functional Block Diagramsp.47System Functional Block Diagramp.47e-box and Power Supply Functional Block Diagramp.49Token Ring Cablep.50Physio Functional Block Diagramp.51Transducer Safety Testing: Test Setup and Theoryp.52Theory of the Transducer Leakage Current Testp.545 Installationp.59Overviewp.59Assemble the Systemp.59Power Cord and Monitor Installationp.59Installing the Power Cord and Monitorp.59Peripherals Installationp.59Installing a VCRp.60Installing a Sony Color Printerp.60Installing a Sony Black-and-White Printerp.61System Startupp.62Powering On the Systemp.62Monitor Signal Voltagep.62Setting the Monitor Signal Input Voltagep.62Printer Preparationp.64Preparing the Sony Black-and-White Printerp.64Preparing the Sony Color Printerp.64System Configurationp.64Localizationp.65Setting the Localization Parametersp.65Options Installationp.71Installing Optionsp.71DICOM Network Configurationp.73Changing the System’s Network Parametersp.73Using the Field Service Options Window for Advanced Network Settingsp.74Adding Server Informationp.77Assigning Servers to DICOM Rolesp.78Configuring Autodeletep.79Configuring Loggingp.81System Hardware Configurationp.83Assigning Record Buttonsp.83Assigning Option Buttonsp.84VCR configurationp.85Configuring the System for PAL Formatp.85Setting the Monitor Color Temperaturep.87Configuring for printersp.87Customer Trainingp.886 Performance Testsp.89Overviewp.89Electrical Safety Testsp.89Chassis to Ground Resistance Testp.90Chassis to Ground Resistance Test Procedurep.91Ground Wire Leakage Current Testp.92Ground Wire Leakage Current Test Procedurep.93ECG Lead Leakage Current Test (Source)p.94ECG Lead Leakage Current Test Procedurep.95ECG Lead Isolation Leakage Current Testp.96ECG Lead Isolation Leakage Current Test Procedurep.97Transducer Testsp.98Transducer Leakage Current Test (Source)p.99Transducer Leakage Current Test Procedurep.100Transducer Isolation Leakage Current Testp.102Transducer Isolation Leakage Current Test Procedurep.103PC Diagnosticsp.105IBM Drive Fitness Testp.105Creating a DFT Bootable Floppy Diskp.105Testing PC Functions Using the Drive Fitness Testp.105Checkit Prop.106Testing PC Functions Using Checkit Prop.106Resident Self Testsp.107Availabilityp.107Logging Onp.107User Interfacep.108Information Typesp.109System Informationp.109Hardware Informationp.110Testsp.111Available Testsp.111Testing Optionsp.112Test Resultsp.113Service Logp.113Launching the RST Applicationp.113Closing the RST Applicationp.114Running a Basic Testp.114Running an Extended Testp.115Test Modulesp.117Running an Interactive Testp.118Viewing Error Logsp.119Ultrasound Application Logp.119Clearing the Ultrasound Application Logp.120RST Test Logp.121RST Log Codesp.122Clearing the RST Test Logp.122Test Result Detailsp.123Viewing Test Detailsp.123System Logp.124Clearing the System Logp.124Using the Service Logp.125Making an Entry in the Service Logp.125Modifying an Entry in the Service Logp.1267 Adjustmentsp.127Monitor Adjustmentp.127OSD Main Menup.127Opening the OSD Main Menup.127Navigating the OSD Menusp.128Adjusting VGA Vertical Sizep.128Adjust the VGA Vertical Sizep.128Changing the Boot Sequencep.130Change the Boot Sequencep.1308 Preventive Maintenancep.131Cleaning the Air Filterp.131Cleaning the Trackballp.132Cleaning the e-boxp.132Degaussing the Monitorp.1339 Troubleshootingp.134PC Won’t Bootp.134PC Boots, but Monitor is Blankp.135PC Boots, But Peripherals Don’t Workp.135Printer Problemsp.136HP940 Startupp.136Printer Installation Problemsp.136Electromagnetic Interferencep.13710 Service Proceduresp.138Overviewp.138Repair Philosophyp.139ESD Safeguardsp.139General Precautionsp.139Service Accessp.140Servicing Components Above the Lower Enclosurep.140Servicing Components in the Lower Enclosurep.140Rear Service Doorsp.141Front Service Panelp.141Side Panelsp.141Record Keepingp.141PC Servicep.142Software Service Proceduresp.142System Softwarep.142Reloading or Upgrading System Softwarep.142Backing Up and Restoring Presetsp.143Backing Up Customer Presetsp.143Restoring Customer Presetsp.146PC Componentsp.147Opening the PC Side Coverp.148Removing the PC from the Systemp.150Eboxp.151Replacing Ebox Circuit Boardsp.151Replacing the Motherboardp.152Replacing the Distribution Boardp.153Power Supplyp.154Removing the Power Supply and the Eboxp.154Separating the Ebox and the Power Supplyp.155Transducer Connector Modulesp.156Replacing Transducer Connector Modulesp.156Fanp.156Replacing the Fanp.156Monitorp.157Replacing the Monitorp.157Monitor Twivelp.158Disassembling the Twivelp.158System Control Panelp.158Replacing the System Control Panelp.159Monitor Support and Front Panelp.160Replacing the Monitor Support and its Front Panelp.160Cablesp.160System Control Panel and Monitor Cablesp.161Replacing the Upper Assembly Cable Bundlep.161Cartp.163Replacing Cart Wheels (Caster Assemblies)p.163Replacing the Top Panel of the Lower Enclosurep.164Lift Columnp.165Preparationp.165Preparing the System for Servicing the Lift Columnp.165Gas Springp.166Replacing the Gas Springp.166Mechlok and Cablep.167Replacing the Mechlok and Cablep.167Slidesp.169Replacing the Slidesp.16911 Cablingp.17112 Configurationp.174Disk Drive Settingsp.174Hard Drive Jumpersp.174CD Drive Jumpersp.175MO Drive Jumpersp.175VCR Switch Settingsp.175Sony B/W Printer Dip Switch Settingsp.17613 Partsp.177Overviewp.177Ordering Partsp.177New Partsp.178Exchange Partsp.178System Serial Number and Revision Levelp.178Major Assembliesp.178Transducer Connector Modulesp.179Internal PCp.180e-box and Power Supplyp.183Lower Assemblyp.185Lift Column Assemblyp.190Upper Assemblyp.193Top Shelfp.196VCR Option Partsp.198Softwarep.198Physio Option Partsp.199Exchange Part Numbersp.200Optional Peripheralsp.201Supplies and Accessoriesp.20214 Transducersp.20515 Glossaryp.206
Page 1 of the Philips Patient Monitor (M2540) user manual — M2540 Ultrasound System Field Service Manual
Page 1 · click to enlarge
Page 1

system, and describes ESD precautions to be taken when servicing the

scan converter sections. This section also includes explanations of the

Section 1, “General Information,” presents a product overview and a

Section 3, “Safety,” discusses safety issues pertinent to the ultrasound

procedures that apply to the M2540A, including electrical safety tests.

safety, and system safety. A procedure for verifying complete system

This manual provides information for troubleshooting and servicing

Section 4, “Theory of Operation,” includes a technical overview of

system functions. Functional descriptions are given for each system

circuit board. Functional block diagrams show the scanner and the

theories behind the electrical safety tests in the Performance Tests

system, how to install peripheral devices, and how to configure the

description of major features, and lists applicable safety standards.

Section 5, “Installation,” explains how to unpack and assemble the

Section 6, “Performance Tests,” comprises all tests and diagnostic

the Philips M2540A ultrasound system. Brief overviews for each

The chapter emphasizes patient, operator, and service personnel

Section 2, “Specifications,” describes the physical and electrical

1 About This Manual

section in this book are listed below:

specifications of the M2540A.

system’s parameters.

operation is included.

system.

section.

About This

Manual

Philips Patient Monitor (M2540) — About This diagram (page 1)
Fig. 1 · About This

Copyright © 2002 by Philips Medical Systems All rights reserved

M2540-92000-01 A November 2002

M2540 Ultrasound System Field Service Manual

Page 2 of the Philips Patient Monitor (M2540) user manual — Audience
Page 2 · click to enlarge
Page 2
Philips Patient Monitor (M2540) — Audience diagram (page 2)
Fig. 2 · Audience
About This ManualAbout This Manual

precautions, component removal and replacement, and peripherals installation. The chapter also

presents software procedures for backup and retrieval of customer presets, and for reloading or

Section 9, “Troubleshooting,” contains information and procedures for finding and repairing the

Section 10, “Service Procedures,” includes all service procedures, including a discussion of ESD

causes of faults in the M2540A, including software error codes. The error-code section opens

Section 12, “Configuration,” lists the jumper and switch settings for the M2540A’s disk drives.

Section 15, “Glossary,” is a glossary of terms used in this manual and in the ultrasound imaging

Section 8, “Preventive Maintenance,” includes procedures for periodic maintenance of the

Audience This manual supports the field service maintenance and repair of the M2540A Ultrasound

System. The user of this document is a qualified ultrasound electronics technician who has

Section 11, “Cabling,” includes tables listing all the standard and optional M2540A cables.

Section 13, “Parts,” comprises parts lists and exploded diagrams of the M2540A and its

Section 14, “Transducers,” lists the transducers compatible with the M2540A and their

Section 7, “Adjustments,” contains instructions for adjusting the system monitor.

completed training classes on the system and its peripherals.

assemblies. Parts ordering methods are discussed.

with a key for interpreting the codes.

upgrading system software.

characteristics.

system.

field.

Page 3 of the Philips Patient Monitor (M2540) user manual — Service Manual
Page 3 · click to enlarge
Page 3
About This ManualAbout This Manual
  • Menu items or titles appearing on the display are spelled and capitalized in the manual as they

If you have questions about the service manual, or if you discover an error in the manual, contact

  • All procedures are numbered. You must complete steps in the sequence they are presented

Acrobat Reader. A list of bookmarks functions as a table of contents. Those bookmarks and

Format This manual is in Portable Document Format (PDF), for viewing on a laptop computer using

  • Bulleted lists indicate general information about a function or a procedure. They do not

cross-references use hypertext links to provide access to the referenced information.

  • Control names are spelled and capitalized in the manual as they are on the system.
  • Technical Publications, MS 964, at the first address below

Conventions The following conventions are used in this manual:

Philips Ultrasound Technical Publications:

  • atl-bothell.techpubs@<a href="http://philips.com" rel="nofollow noopener" target="_blank">philips.com</a>
  • An English system is assumed.

imply a sequential procedure.

  • Hypertext links are blue .

to ensure reliable results.

are on the display.

Service Manual

Questions or

Comments

Page 4 of the Philips Patient Monitor (M2540) user manual — USA
Page 4 · click to enlarge
Page 4
About This ManualAbout This Manual

Non-Philips Medical Systems product names may be trademarks or registered trademarks of their respective owners.

Assistance Various support locations around the world can provide customers with technical assistance

regarding the ultrasound system. Customers should contact the sales office where they

purchased the system or the nearest Philips Ultrasound office for assistance.

To find your local service center phone number, go to:

Philips Medizin Systeme Boeblingen GmbH

3000 Minuteman Road Andover, Massachusetts 01810-1099

(425) 487-7000 or (800) 426-2670

  • Authorized EU Representative:

www.ultrasound.philips.com

Hewlett-Packard-Str. 2

www.philips.medical.com

  • Philips Ultrasound
  • Philips Ultrasound

P.O. Box 3003

USA

Customer

Bothell, WA 98041-3003 · 71034 Boeblingen Germany · (978) 687-1501

Page 5 of the Philips Patient Monitor (M2540) user manual — M2540-98000-01
Page 5 · click to enlarge
Page 5
About This ManualAbout This Manual

This system complies with the

0123 marking is for Council Directive

Medical Device Directive.

Representative:

93/42/EEC.

The cautions explain potential

other symbols appearing on the

the user documentation for that

system complies with the FDA

this device to use by or on the

This symbol on the

This symbol on the

The United States Food and

Symbols on the System

electrical shock.

The monitor used in this

applicable at the date of

damage to equipment.

system or its probes:

CAUTION

part of the system.

Subcategory J).

statement:

Text:

!

only supplied power cords and

Do not operate the system in the

transducer for the first time, be

and Use” section of the chapter

explosion hazard, the safety of

Do not remove system covers.

To avoid electrical shock, use

grounded (3-hole) wall outlets.

sure to read the “Description

transducer. Also, for TEE and

review the “Electrical Safety”

guidelines for setting controls

that affect acoustic output and

Pay special attention to the

guidelines for fetal use, and

Before you use a specialty

sections in those chapters.

The warnings explain the

connect only to properly

that is applicable to your

WARNING

measurements.

WARNING

anesthetics.

Text:

© 2002 Philips North America

Reproduction in whole or part

is prohibited without the prior

written consent of the copyright

Philips Medical Systems makes

including, but not limited to, the

merchantability and fitness for a

Philips Medical Systems shall

damages in connection with the

Windows XP is a trademark of

The information contained in

no warranty of any kind with

PublishedDecember 2002

this document is subject to

change without notice.

regard to this material,

not be liable for errors

contained herein or for

Cannon is a registered

All Rights Reserved.

or use of this material.

M2540-98000-01

Printed in U.S.A.

particular purpose.

Corporation.

holder.

Show 32 more labels / short text on this page
71034 Boeblingen, Germany · Hewlett-Packard Strasse 2 · Philips Medizinsysteme · Boeblingen GmbH · Authorized EU · Drug Administration requires · operators to consult · system indicates a · system advises · potential for · the following labeling · regulations that were · manufacture (21 CFR · Prescription Device · Monitor Radiation · Electrical Shock Hazard · intraoperative transducers, · ultrasound, applications, · presence of flammable · Safety Information · accuracy of clinical · Explosion Hazard · incidental or consequential · furnishing, performance, · implied warranties or · trademark of Cannon · Microsoft Corporation · Publication number · Corporation · Trademarks · Warranty · Edition 1
Page 6 of the Philips Patient Monitor (M2540) user manual — Overview
Page 6 · click to enlarge
Page 6
General InformationProduct Overview

The M2540A ultrasound system comprises a cart, a system control panel, and a monitor. A foot

pedal on each caster locks and unlocks the front cart wheels. All four wheels swivel. The system

control panel and the monitor adjust up and down and swivel as a unit, and the monitor tilts and

Most peripheral devices mount on top of the lower enclosure. The optional plain-paper printer

The Resident Self Test (RST) software verifies system performance and helps diagnose problems.

foot switch, and a magneto-optical disk (MOD) drive. The VCR is available in either NTSC or

PAL configuration. For a complete list of optional components, see “Optional Peripherals” on

Optional peripheral components include a video cassette recorder, various types of printer, a

The cart’s lower enclosure contains the M2540A’s computer, disk drives, and ultrasound

is not mounted on or powered by the system, but is connected from a remote location.

  • “Optional Peripherals (VCR, Printers, and Image Devices)” on page 9

CD-RW and floppy disk drives are standard equipment.

This chapter includes the following sections:

1 General Information

  • “Safety Specifications” on page 12
  • “Product Overview” on page 6
  • “System Description” on page 7
  • “Preset Functions” on page 10

generating and processing boards.

swivels on its mount.

page 201.

Overview

Product

Page 7 of the Philips Patient Monitor (M2540) user manual — Cart Wheels
Page 7 · click to enlarge
Page 7
General InformationSystem Description

include important features of the cart, monitor, system control panel, Physio module, e-box, and

The panel is at the center of the lower part of the cart, and is accessible to the system operator.

Four wheels at the cart base provide system maneuverability and braking. All four wheels swivel,

and unlock the front wheels to prevent the cart from rolling. Locking the wheels immobilizes the

The system control panel includes four numbered option buttons whose functions are assigned

Major ultrasound system components are described on the following pages. These descriptions

References to the left and the right sides of the system are as viewed from the front of the cart.

A height adjustment lets you adjust the system control panel and monitor for operator comfort.

communication between the system and a network, and for controlling certain legacy printers.

The foot switch port, an equipotential lug, and the system ground lug are also on the I/O panel.

and the front wheels can be fixed straight and locked. Foot pedals on the two front wheels lock

components as standard equipment. These include the acoustic processor input/output (APIO)

Cart The cart supports the system and acts as the chassis into which all system components are

PC The M2540A uses a personal computer (PC) as a central processor. The PC houses several

by the customer. See the online Help file for details on configuring these controls.

An Input/Output (I/O) panel at the rear of the cart contains three connectors for

See “I/O Panel” on page 45 for more information about the I/O panel.

board, a video card, a CD-RW drive, and a floppy disk drive.

system during patient procedures.

installed and interconnected.

system power supply.

Cart Wheels

Description

System

Page 8 of the Philips Patient Monitor (M2540) user manual — System Control
Page 8 · click to enlarge
Page 8
General InformationSystem Description

magneto-optical disk (MOD) drive can also be installed in one of the drive bays. The VCR option

Panel The system control panel at the top front of the cart is a replaceable, self-contained module. See

“Replacing the System Control Panel” on page 159. The panel interfaces with the imaging system

through a USB cable that connects to the internal PC. Power for the trackball is provided via the

USB cable. All the other control-panel functions are powered by a cable from the system power

See “Internal PC” on page 39 for more detailed information about the PC and its components.

e-box The e-box houses the scanner circuit boards, and is accessible by opening the door at the right

System Monitor The monitor at the top of the cart is a 15-inch color display mounted on a “twivel” assembly.

includes three boards that reside in the PC’s PCI slots, and a VCR that mounts on the cart.

The system control panel contains a backlit alphanumeric keyboard, slide controls, rotary

The optional Physio module installs in one of the PC’s front drive bays. An optional

The twivel allows tilt and swivel positioning of the display for ease of viewing.

controls, hard-coded and software-driven keys, and a trackball.

  • A beam processor/acoustic processor (BPAP) board

There are several circuit boards in the e-box:

  • Two transmit and receive (TR) boards
  • The system motherboard
  • The distribution board
  • A demodulator board

rear of the cart.

supply.

System Control

Page 9 of the Philips Patient Monitor (M2540) user manual — Printers, and
Page 9 · click to enlarge
Page 9
General InformationOptional Peripherals (VCR, Printers, and Image Devices)

e-box at the right side of the cart. It accepts AC input sources from 100 V to 240 V, at 50 Hz or

120 VAC outlets for powering the monitor and two peripheral devices are at the rear of the power supply. An unswitched outlet powers the M2540A’s internal PC. When the system is shut

down, only the 120 VAC outlet to the PC is enabled. A green LED on the left front of the power

Physio Module The optional Physio module (see “Physio Module Option” on page 43) installs in one of the PC

An externally generated auxiliary analog input signal using the phone jack can substitute for the

3-lead ECG physio connector at the front panel. You can also input an auxiliary ECG waveform from an external patient monitor and display it, in all modes. The R-wave from this signal can

“Optional Peripherals” on page 201 lists all peripherals supported by the ultrasound system.

Power Supply The power supply is an enclosed, self-contained, replaceable module that mounts under the

The power supply provides all required AC and DC voltages to the system. Three switched

drive bays. The front panel of the module bears two input connectors:

  • Various biopsy kits (See “Supplies and Accessories” on page 202 .)

The M2540A supports several types of optional peripheral devices:

Some peripherals mount on a shelf below the system control panel.

  • An auxiliary analog input (standard ¼-inch phone jack)
  • VCRs, including NTSC-format and PAL-format models
  • A magneto-optical disk (MOD) drive for data storage
  • Thermal color and thermal black-and-white printers

supply lights when the power supply is energized.

serve as a time reference for the system.

  • A 3-lead ECG physio connector

60 Hz.

Printers, and

Peripherals

Optional

Devices)

(VCR,

Image

Page 10 of the Philips Patient Monitor (M2540) user manual — Functions
Page 10 · click to enlarge
Page 10
General InformationPreset Functions

The ultrasound system includes programmable presets that configure the system for best imaging

preferred configurations. If preset configurations are subsequently changed, they can be

quickly restored from the backup copy, without having to reset them manually. Keeping

For specific information regarding the configuration and use of presets or other features of the

results in a particular situation. Activating a preset initializes system settings to values that are

optimal for a chosen exam. The factory installed software includes several presets, tailored to

different applications. The user can define as many as 20 additional presets in each of 8 exam

a backup copy also eliminates the need to manually reconfigure the presets after a

types, to adjust system variables (including acoustic power) to any required configuration.

software upgrade. See “Backing Up and Restoring Presets” on page 143 for the

procedures to back up and restore presets.

system, see the online Help file.

Functions

Preset

Page 11 of the Philips Patient Monitor (M2540) user manual — Equivalencies
Page 11 · click to enlarge
Page 11
General InformationKeyboard Equivalencies

Some of the keys found on a standard PC keyboard are not present on the M2540A. Many of the

keyboard. The M2540 keys act as PC-equivalent keys whenever the ultrasound application is not

running. The following table lists the missing PC keys and their equivalents on the M2540A:

functions performed by those keys are mapped to keys that are present on the M2540A

PC Key M2540A Key PC Key M2540A Key

F1 Patient F9 Blank 1

F2 Preset F10 Blank 2

F3 Review F11 Blank 3

F4 Report F12 Probe

F6 Help Delete Fusion

F8 Mic Page Down Right

F5 Setup Esc THI

F7 VCR Page Up Left

Table 1-1 Keyboard Equivalents

Equivalencies

Keyboard

Page 12 of the Philips Patient Monitor (M2540) user manual — Safety Limits
Page 12 · click to enlarge
Page 12
General InformationSafety Specifications

Ground wire leakage See Section 6, “Performance Tests”, and the Safety and Standards Guide

Wheel locks With wheel locks engaged, system remains stationary on slopes of up

External sharp edges The system exterior has no sharp edges, in compliance with IEC

Compliance For information about applicable safety standards and specifications, consult the Safety and

System tip over Will not tip over on an incline of up to 10° under normal use.

temperature System surfaces do not exceed temperature limits specified in

components The power supply and monitor comply with UL 2601-1 1997.

This section describes the safety specifications of the M2540A.

IEC 60601-1 and EN 60601-1

Power supply Complies with IEC60601-1

shipped with the M2540A.

to 5 ° in any orientation.

60601-1 and EN 60601-1

Standards Guide shipped with the M2540A.

Table 1-2 Safety Limits

Specifications

Safety Limits

Regulatory

Safety

Parameter Specification · Potentially hazardous · System surface

Page 13 of the Philips Patient Monitor (M2540) user manual — 2 Specifications
Page 13 · click to enlarge
Page 13

Specifications:

This section lists the specifications of the M2540A ultrasound system. The following specification

  • “Connection and Communication Specifications” on page 21
  • “Audio/Video Specifications” on page 23
  • “Transducer Specifications” on page 20
  • “Physio Port Specifications” on page 23
  • “Electrical Specifications” on page 19
  • “Physical Dimensions” on page 14

types are included in this section:

  • “Monitor” on page 19

2 Specifications

Page 14 of the Philips Patient Monitor (M2540) user manual — Dimensions
Page 14 · click to enlarge
Page 14
SpecificationsPhysical Dimensions

Depth (front to back) 103 cm (40.5 inches) See Figure 2-2

Width 53.5 cm (21 inches) See Figure 2-1

Highest position147.5 cm (58 inches) See Figure 2-1

Height of lower enclosure 76.2 cm (30 inches) See Figure 2-2

Monitor to control panel front edge 43.18 cm (17 inches) See Figure 2-2

(front to back) 41.71 cm (16.42 inches) See Figure 2-2

Weight Less than 91 kg (200 lbs) including display but no peripherals See Figure 2-2

The tables and diagrams that follow detail the physical specifications of the M2540A.

(VCR and printers temperature limit0° to 40° C at 80% RH)
Relative humidity20% to 90%, non-condensing
Atmospheric pressure572hPa to 1013 hPa
Atmospheric pressure572hPa to 1013 hPa

Height (to monitor top) Lowest position: 129.5 cm (51 inches)

Temperature range–40° to 55° C
Temperature range0° to 40° C
Relative humidity20% to 80%

Table 2-1 Physical Specifications

Dimensions

Physical

Show 5 more labels / short text on this page
Parameter Specification Reference · Environmental Operational · Storage · Monitor depth · Dimensions
Page 15 of the Philips Patient Monitor (M2540) user manual — Physical Dimensions
Page 15 · click to enlarge
Page 15
SpecificationsPhysical Dimensions
Philips Patient Monitor (M2540) — Physical Dimensions diagram (page 15)
Fig. 3 · Physical Dimensions

129.5 cm (51 in) - 147.5 cm (58 in)

Figure 2-1 System Front Dimensions

Page 16 of the Philips Patient Monitor (M2540) user manual — Physical Dimensions
Page 16 · click to enlarge
Page 16
SpecificationsPhysical Dimensions
Philips Patient Monitor (M2540) — Physical Dimensions diagram (page 16)
Fig. 4 · Physical Dimensions

Figure 2-2 System Side Dimensions

Page 17 of the Philips Patient Monitor (M2540) user manual — Physical Dimensions
Page 17 · click to enlarge
Page 17
SpecificationsPhysical Dimensions
Philips Patient Monitor (M2540) — Physical Dimensions diagram (page 17)
Fig. 5 · Physical Dimensions

129.5 cm (51 in) - 147.5 cm (58 in)

Figure 2-3 System Rear Dimensions

Page 18 of the Philips Patient Monitor (M2540) user manual — Physical Dimensions
Page 18 · click to enlarge
Page 18
SpecificationsPhysical Dimensions
Philips Patient Monitor (M2540) — Physical Dimensions diagram (page 18)
Fig. 6 · Physical Dimensions

Figure 2-4 System Top Dimensions

Page 19 of the Philips Patient Monitor (M2540) user manual — Monitor
Page 19 · click to enlarge
Page 19
SpecificationsElectrical Specifications

AC input 90 VAC to 264 VAC, 47 Hz to 63 Hz Ground impedance 200 milliohm maximum Dielectric withstand 1500 VAC mains to safety ground 2000 V mains to AC secondaries 4000 V mains to DC secondaries Load 1150 VA maximum. AC output 120 VAC, 60 Hz, quasi square-wave, 500 VA maximum

Table 2-2 Electrical Specifications

75 Hz refresh rate RGB color display

Screen size 15-inch diagonal Display format VGA, 800x600

Table 2-3 Main Display

Features Tilt ±30 degrees

Monitor

Specifications

Electrical

Swivel ±135 degrees · Parameter Specification · Parameter Specification

Page 20 of the Philips Patient Monitor (M2540) user manual — Specifications
Page 20 · click to enlarge
Page 20
SpecificationsTransducer Specifications

The following table lists the specifications of the transducers that the M2540A supports.

PA 4-2 biopsy capable 21422A 2 to 4 MHz Advance Vision

CA 5-2 21425A 2 to 5 MHz Advance Vision

1. This transducer is compatible with the M2540A, but is only sold separately.

S4 1 21330A Cannon

c3540 biopsy capable 1 21321A 3.5 MHz Cannon

EC6509 endocavity 21336A Cannon

L7535 linear 21359A 7.5 MHz Cannon

TEE (Omni II) 21369A 6 MHz Cannon

L1038A small parts 21376A Cannon

15-6L intraoperative 21390A Cannon

D1914c 21221B 1.9 MHz Pencil

Specifications

Transducer

Show 5 more labels / short text on this page
Number Frequency Connector · S8 21350A Cannon · S12 21380A Cannon · Table 2-4 Transducers · Description Part
Page 21 of the Philips Patient Monitor (M2540) user manual — Ports PC Ports
Page 21 · click to enlarge
Page 21
SpecificationsConnection and Communication Specifications

Monitor 15-pin D connector on graphics adapter board Foot switch 9-pin female D-sub connector on the APIO board that carries foot switch signals from the external I/O panel

The following table lists the ports on the rear of the internal PC and describes their locations.

Speakers Green 3.5 mm stereo phone jack to system speaker amplifier

The following tables list the communication and connector specifications of the M2540A.

(optional) 1 Female phono (RCA) on TV and video converter board

S-Video output (optional) 1 4-pin mini circular DIN on SVGA to TV video card S-Video input (optional) 1 4-pin mini circular DIN on video capture card Print trigger output 3.5-mm phone jack on APIO board VCR audio output (line out)

(optional) 1 Green 3.5-mm stereo phone jack on sound card

VCR audio input (line in) 1 Blue 3.5-mm stereo phone jack on sound card

Microphone Pink 3.5-mm stereo phone jack

Table 2-5 PC Ports

Ports PC Ports

and Commu-

Specifications

Connection

System I/O

nication

Port Specification · Composite video output

Page 22 of the Philips Patient Monitor (M2540) user manual — Connection and Communication Specifications
Page 22 · click to enlarge
Page 22
SpecificationsConnection and Communication Specifications

Com 1 9-pin D connector, RS-232; spare serial port used for serial VCRs

USB USB (A) for data and power to system control panel

Token ring 50-pin connector on the APIO board LAN RJ-45 to I/O panel for network communication

USB For black-and-white thermal printer

USB For remote plain-paper printer

1. Only available with VCR option or external video/print option

Keyboard DIN circular connector

Mouse DIN circular connector

Table 2-5 PC Ports (Continued)

USB For color printer

PS-2 keyboard Not used

PS-2 mouse Not used

Parallel port Not used

Audio input on main board Not used

Port Specification

Page 23 of the Philips Patient Monitor (M2540) user manual — Specifications
Page 23 · click to enlarge
Page 23
SpecificationsPhysio Port Specifications

Each signal is a contact closure to ground, active low, and

Foot switch D-sub 9 Control signal from foot switch to APIO board

Composite video BNC Composite video signal from SVGA board

I/O Panel Ports The following table lists the ports on the I/O panel and the signals they carry.

Defaults arerecord, freeze, and record 2

LAN RJ-45 Communication with local area network

The following tables list the audio and video specifications of the M2540A.

The following table lists the specifications of the M2540A’s physio ports.

Sensitivity3 mVpp ±2.5 mV for full scale at 100% gain

Print BNC Trigger signal from APIO board

Aux 1 3-dB bandwidthfull scale DC to 100 Hz minimum Maximum input signal: ±4 V
Sensitivity2.5 ±0.5 Vpp for full scale at 100% gain

Table 2-7 Physio Port Specifications

Frequency response1 ±0.5 to 30 ±6 Hz

ECG 3 patient leads with R-wave detection Monitoring quality only

Table 2-6 I/O Panel Ports

Specifications

Audio/Video

Specifications

Physio Port

TTL compatible · Port Connector Signal · Port Specification

Page 24 of the Philips Patient Monitor (M2540) user manual — External Video
Page 24 · click to enlarge
Page 24
SpecificationsAudio/Video Specifications
Active lowON = 0.5 V max @ 1 mA; OFF = 5.25 VDC maximum

Doppler spectrum away from transducer — right speaker

1.0 V p to p ±5% into 75 ohms (only with VCR option) Print Trigger 1 signal from the APIO board

Doppler spectrum toward transducer — left speaker

Microphone Faces front. 10 Hz to 20 KHz electret type

Table 2-9 External Video Specifications

Color Composite NTSC 3.58 (USA), PAL 4.43 (Europe)

Maximum power input4 watts

Table 2-8 Audio Specifications

Impedance8 ohms

Response 130 Hz to 12 KHz

External Video

Audio

Show 5 more labels / short text on this page
Speakers Stereo—VCR, Doppler · Description Specification · Description Specification · voltage · Frequency
Page 25 of the Philips Patient Monitor (M2540) user manual — Video Output
Page 25 · click to enlarge
Page 25
SpecificationsAudio/Video Specifications

VCR 1.0 Vpp luminance incorporating sync and 0.3 Vpp chrominance as shown in Table 2-9. This timing is essentially RS-170 (60 Hz) or CCIR

Table 2-10 Video Output Specifications

monitor 0.7 Vpp RGB with TTL sync

(50Hz) video timing.

Video Output

Specifications

Output Specification · RGBSync to system · S-Video Output to

Page 26 of the Philips Patient Monitor (M2540) user manual — 3 Safety
Page 26 · click to enlarge
Page 26

Safety:

This section provides an overview of safety considerations for the ultrasound system. These

safety concerns apply to patients, operators, and service engineers. For more detailed safety

  • “AIUM/NEMA Output Display Standard” on page 28

information, see the Philips Safety and Standards Guide.

The following topics are included in this section:

  • “Electromagnetic Compatibility” on page 35
  • “Transmit Power (Acoustical)” on page 27
  • “Glutaraldehyde Exposure” on page 32
  • “Peripheral Connections” on page 31
  • “Restrictions for Use” on page 36
  • “Electrosurgical Units” on page 37
  • “Acoustic Exposure” on page 27
  • “Electrical Warnings” on page 31
  • “Moving the System” on page 32
  • “Explosive Hazards” on page 30

3 Safety

Page 27 of the Philips Patient Monitor (M2540) user manual — NOTES
Page 27 · click to enlarge
Page 27
SafetyTransmit Power (Acoustical)

diagnostic image with minimal power output. A display standard presents this index using one of

  • The power index setting on the System Setup menu selects any of the four power indices
  • The Power rotary control at the right side of the system control panel is the only control

For additional information on acoustical power settings and power index, see the Philips Safety

Acoustic output, expressed as an index, is displayed on the screen to allow the best possible

The index setting on the System Setup menu selects the power index used. The displayed

Although no harmful effects have been demonstrated for any of the ultrasound frequencies,

intensities, and exposure times used in examinations with Philips ultrasound systems, Philips

recommends that you consider the following, and use the lowest ultrasound exposure that

index is based on this setting and on preset configuration and imaging mode.

  • Use diagnostic ultrasound only when there is a good medical reason.
  • Reset the controls at the start of every examination.

produces diagnostically acceptable information:

  • Thermal index for cranial bone (TIC)
  • Thermal index for soft tissue (TIS)

that affects transmit power level.

the following four power indices:

  • Thermal index for bone (TIB)

for display at any time.

  • Mechanical index (MI)

and Standards Guide.

NOTES

(Acoustical)

Transmit

Acoustic

Exposure

Power

Page 28 of the Philips Patient Monitor (M2540) user manual — MI: 0.8
Page 28 · click to enlarge
Page 28
SafetyAIUM/NEMA Output Display Standard
  • Select a transducer that provides good resolution and focal depth for the region of interest.

For more detailed information on acoustic exposure, see the Output Display Standards and ODS

type of index displayed and the value of that index for the acoustic output currently being used.

The thermal index for bone (TIB) is used in applications such as second or third trimester fetal

Institute of Ultrasound in Medicine and the National Electrical Manufacturers Association, the

Philips ultrasound system displays power output indices related to the potential for bioeffects.

For example, if the output corresponds to a mechanical index of 0.8, the following is displayed:

The displayed index is one of four types: MI, TIS, TIB, or TIC. These ultrasound abbreviations

  • Use techniques that let you collect clinical data quickly and end the examination promptly.

Real-time information related to the power output is displayed on the monitor, indicating the

In compliance with the Output Display Standard (ODS) jointly proposed by the American

Soft tissue thermal index (TIS) is used in cardiac, fetal, and abdominal scanning.

The cranial bone thermal index (TIC) is used for transcranial imaging.

scanning and neonatal cephalic (through the fontanelle) scanning.

  • Reduce exposure time, independent of acoustic index value.

Then use the imaging controls to fine-tune image resolution.

conform to the AIUM/NEMA Output Display Standard.

Acoustic Tables booklet.

MI: 0.8

AIUM/NEMA

Standard

Output

Display

Page 29 of the Philips Patient Monitor (M2540) user manual — Automatic
Page 29 · click to enlarge
Page 29
SafetyAIUM/NEMA Output Display Standard

If none of the conditions in “When MI is Displayed With the Normal Setting” on page 29 exist, if

power indices is selectable for display at any time, using the power index setting in the

index setting of Normal from the System Setup menu. This directs the system to choose an

Index Selection For automatic selection of a power index based on system mode, the user selects the power

transducer type, the imaging mode, and the selected power index. Any of the four

no transcranial preset is active, and any of the following conditions exist, TIS displays:

  • Black-and White Doppler Preview is the active imaging mode.
  • Doppler Spectral is the active imaging mode (with 2D Live off).
  • Black-and White MMode Preview is the active imaging mode.
  • Black-and White Doppler 2D Live is the active imaging mode.

When MI is Displayed With the Normal Setting

When TIS is Displayed With Normal Setting

MI is displayed if any of the following conditions exist:

index based on the active preset and imaging mode.

  • MMode Trace is the active imaging mode.
  • 2D Only is the active imaging mode.

System Setup menu.

  • Color is turned on.
  • Angio is turned on.

Automatic

Page 30 of the Philips Patient Monitor (M2540) user manual — Explosive
Page 30 · click to enlarge
Page 30
SafetyExplosive Hazards

If none of the conditions in “When MI is Displayed With the Normal Setting” on page 29 exist, a

The displayed index value does not provide an exact value of the potential for adverse bioeffects

in the patient. However, for any patient, the higher the value, the higher the potential for adverse

bioeffects. The user can minimize the potential for bioeffects by keeping the index value as low as

possible, by choosing the right transducer and making adjustments. Minimizing examination time

preferred acoustic power display format, regardless of the mode, the transducer, or the preset

If the user selects an ODS setting other than Normal, the selected index type is used as the

likely that a particular patient's actual conditions are better than indicated by the index. The

transcranial preset is active, and any of the following conditions exist, TIS displays:

operator should be aware of patient conditions that mitigate the actual exposure.

  • Do not operate the system in the presence of flammable anesthetics.
  • Doppler Spectral is the active imaging mode (with 2D Live off).

When TIC is Displayed With Normal Setting

Observe the following practices to avoid explosive hazards:

  • MMode Trace is the active imaging mode.

also minimizes bioeffects.

  • Color is turned on.
  • Angio is turned on.

that is selected.

Explosive

Hazards

Page 31 of the Philips Patient Monitor (M2540) user manual — Connections
Page 31 · click to enlarge
Page 31
SafetyElectrical Warnings

power outlet on the imaging system, or from an isolation transformer that meets medical safety

standards. The 120 VAC outlets on the power supply are isolated. The specific peripherals listed

off. Be sure system power is on before the operation starts, and leave it on for the duration

  • When using the imaging system in the operating room, do not switch system power on or
  • Only qualified service personnel should remove system covers (trim and service panels).

peripherals within six feet of a patient unless the peripherals receive power from an isolated

  • Use only the power cords supplied with the system, and connect them only to properly
  • Do not connect the ultrasound system to the same circuit used for life support devices.

Accidental contact with electrical circuits inside the system could cause serious injury.

in “Optional Peripherals” on page 201 meet medical device standards when installed in the

requirements, but do not meet medical device standards. Therefore, do not use system

Peripherals (such as a VCR or a printer) typically meet general electrical safety usage

Observe the following precautions to prevent electric shock:

grounded electrical outlets.

system as recommended.

of the procedure.

Connections

Peripheral

Electrical

Warnings

Page 32 of the Philips Patient Monitor (M2540) user manual — Before Moving
Page 32 · click to enlarge
Page 32
SafetyGlutaraldehyde Exposure

regulation dealing with levels of acceptable glutaraldehyde exposure in the working environment.

the system, since the ability to move any ultrasound machine is directly related to an individual’s

  • Before transporting the system in a vehicle, remove the monitor and all transducers from the

Philips does not sell glutaraldehyde-based disinfectants with its products. However, this type of

The ultrasound system has been designed to be as lightweight and mobile as possible. However,

approximately 100 kg (220 pounds). Because of this weight, you must use caution when moving

  • Before moving the system, be sure to remove any loose equipment from the top of the sys-

injured their backs moving similar systems. These complaints could not be directly tied to one

particular incident. They do, however, point out the need to be careful in transporting medical

To minimize exposure to glutaraldehyde fumes, make sure the area is well ventilated and use

Although the system is designed to be mobile, remember that it is very heavy, and that you

Some sonographers, particularly those weighing less than 100 pounds, have stated that they

The United States Occupational Safety and Health Administration (OSHA) has issued a

disinfectant is recommended for disinfection of transesophageal (TEE) or endocavity

the system weight—including the weight of the monitor, a printer, and a +VCR—is

tem, disconnect the system power cord, and disconnect all external devices.

the System To move the system, take the precautions listed in the following sections, and

must take precautions when moving the system.

equipment such as an ultrasound system.

system and put them in a packing box.

appropriate eye and skin protection.

size and strength.

transducers.

Before Moving

Moving the

Glutaral-

Exposure

dehyde

System

Page 33 of the Philips Patient Monitor (M2540) user manual — Uneven Surfaces
Page 33 · click to enlarge
Page 33
SafetyMoving the System
5. After the system is in position, engage the wheel locks to immobilize the system, and unlock the control top to allow it to pivot.

Do not move the system over uneven elevator entrances by lifting the machine or any part of the

feet or steeper than 5 degrees. (Wheelchair ramps are usually less than 5 degrees.) Avoid ramps

Always use two people when moving the ultrasound system up and down ramps longer than 20

protocol, the system will not tip over on an incline of up to 10 degrees in any direction. When

System Tilting The system has been tested for stability using the IEC 60601-1 test protocol. Following this

Always use the handle at the front of the cart to move the system from place to place.

that are steeper than 10 degrees, to eliminate the danger of the system tipping over.

this amount of incline is exceeded, there is a potential for the system to tip over.

2. Make sure the control top is locked, to prevent its pivoting during transport.

the System This system is equipped with a front handle and brakes on the front wheels.

3. Engage the track locks on the front wheels to ease straight-line travel.
1. Unlock the wheel locks before moving the system.
4. Push with the handle at the front of the cart.

machine.

Uneven Surfaces

When Moving

Moving on

Ramps or

Page 34 of the Philips Patient Monitor (M2540) user manual — the System in a
Page 34 · click to enlarge
Page 34
SafetyMoving the System

The system’s weight on an extended loading lift may cause the transporting vehicle to tilt, which

Be sure that the transporting vehicle can handle the weight of the system (or systems) plus the

chocks, restraining straps, or other similar types of constraints as an added safety measure. Do

Always engage the wheel locks while transporting the system in a vehicle, and use restraining

Always secure the ultrasound system while it is on the loading lift so that it cannot roll. Make

sure the control top is locked, to prevent its pivoting. Engage the wheel locks and use wood

straps to secure the system in place. Do not rely on the wheel locks to hold the system on

Be sure the load capacity of the loading lift can accommodate the weight of the ultrasound

Load and unload the ultrasound system while the transporting vehicle is parked on a level

surface. The system’s weight can easily cause it to roll on any incline.

system. A minimum capacity of 550 pounds is recommended.

can cause the system to exceed 10 degrees of incline.

not attempt to hold the system in place manually.

could cause personal injury or system damage.

inclines greater than 5 degrees.

passengers.

the System in a

Transporting

Vehicle

Page 35 of the Philips Patient Monitor (M2540) user manual — Electromag-
Page 35 · click to enlarge
Page 35
SafetyElectromagnetic Compatibility

Electrotechnical Commission (IEC 60601-1-2). This IEC standard has been adopted in Europe as

Medical devices can generate or receive electromagnetic interference (EMI). The EMC standards

susceptible to EMI generated by other RF energy sources. Examples of other sources of EMI are

medical devices, information technology products, and radio and television transmission towers.

describe tests for both emitted and received interference. Emission tests deal with interference

Ultrasound systems are designed to receive radio frequency (RF) energy and are therefore

Only a physician can determine if an artifact caused by radiated interference has a negative

Be sure the ultrasound system is firmly secured while inside the transporting vehicle. Any

international standard for EMC with medical devices, as determined by the International

movement, combined with the system’s weight, could cause the system to break loose.

This system has been tested for electromagnetic compatibility (EMC) according to the

generated by the device being tested. The Philips ultrasound system does not generate

See “Electromagnetic Interference” on page 137 to identify sources of EMI.

Tracing the source of radiated interference can be a difficult task.

transporting precautions listed in the sections above.

interference based on the tests described in the standards.

impact on image quality and the subsequent diagnosis.

the European Norm EN 60601-1-2.

exceed the lift's load capacity.

Compatibility

Electromag-

Electromag-

Interference

Avoiding

netic

netic

Page 36 of the Philips Patient Monitor (M2540) user manual — Immunity Level
Page 36 · click to enlarge
Page 36
SafetyRestrictions for Use

first sign of an artifact is seen. This method has two advantages: It removes the issue of subjective

The criteria used for measuring degradation are not specified by the standard, and can vary with

Discharge Electrostatic discharges can cause the ECG heart rate display to increase by 10 to 15% for a few

Immunity is defined in the standard as the ability of a system to perform without degradation in

seconds after the discharge. However, the ECG heart rate display returns to normal within four

assessment that can be subjective. The simplest way to assess degradation is to note when the

receive and amplify low-level signals in the same bandwidth as the interference to which it is

frequencies. This testing showed PW Doppler to be the most susceptible to radio-frequency

the presence of an electromagnetic disturbance. Degradation in image quality is a qualitative

The M2540A ultrasound system is subject to certain restrictions that are described in the

electrostatic discharge immunity levels for their systems. This type of device is designed to

Philips has tested each class of transducer for every operating mode over a wide range of

Please see the Philips Safety and Standards Guide shipped with your system for additional

Test Results The EMC standards require that manufacturers of patient-coupled equipment specify

decision-making and provides the most stringent test results.

information about compliance with EMC standards.

following sections.

the manufacturer.

interference.

susceptible.

seconds.

Immunity Level

Restrictions

Electrostatic

for Use

Page 37 of the Philips Patient Monitor (M2540) user manual — Electrosur-
Page 37 · click to enlarge
Page 37
SafetyElectrosurgical Units

electromagnetic fields or currents into patients. Because imaging ultrasound frequencies are also

in the radio frequency range, ultrasound transducer circuits are susceptible to radio frequency

Electrosurgical units (ESUs) and other devices intentionally introduce radio frequency

interference. While an ESU is in use, the noise generated severely interferes with the

black-and-white image and completely obliterates the color image.

Electrosur-

gical Units

Page 38 of the Philips Patient Monitor (M2540) user manual — 4 Theory of Operation
Page 38 · click to enlarge
Page 38
Theory of OperationOverview

The M2540A’s internal PC replaces the back-end electronics used in earlier ultrasound systems

to handle display processing. The beam processor/acoustic processor (BPAP) board handles the

Overview This section provides a technical overview of system functions. The discussion focuses on the

main functions and features of the system’s PC, circuit boards, and power distribution. Main

system functions are also shown in the functional block diagrams at the end of this section.

  • “Transducer Safety Testing: Test Setup and Theory” on page 52
  • “Distribution Board and Connector Modules” on page 44

This section covers the following areas of the system:

acoustic and beam processing front-end functions.

  • “Functional Block Diagrams” on page 47

4 Theory of Operation

  • “Physio Module Option” on page 43
  • “System Motherboard” on page 44
  • “System Control Panel” on page 45
  • “Demodulator Board” on page 43
  • “Power Supply” on page 46
  • “APIO Board” on page 43
  • “Internal PC” on page 39
  • “BPAP Board” on page 42
  • “TR Boards” on page 40
  • “I/O Panel” on page 45
Page 39 of the Philips Patient Monitor (M2540) user manual — PC Ports
Page 39 · click to enlarge
Page 39
Theory of OperationInternal PC

Internal PC The M2540A uses an internal PC to coordinate and perform many of the functions that required

dedicated circuit boards in earlier ultrasound systems. The power button on the front of the PC

  • The Acoustic Processor I/O (APIO) board, in a PCI slot (See “APIO Board” on page 43 )
  • A USB port communicates with an optional black-and-white thermal printer.

is the system power switch. Turning on the PC starts the ultrasound system.

  • A 3.5 mm phono jack supplies the signal that drives the system’s speakers.

Components The M2540A’s internal PC contains the following standard components:

  • A USB port communicates with an optional plain-paper printer.
  • A 3.5 mm phono jack receives audio signals from a microphone.
  • A USB port communicates with the optional physio module.
  • A USB port communicates with the system control panel.
  • A USB port communicates with an optional color printer.
  • An RS-232 port (Com1) controls an optional VCR.

The PC motherboard’s ports are used as follows:

  • A LAN port connects the PC to a network.
  • A graphics card in the AGP slot
  • A 3 1 / 2 -inch floppy disk drive
  • A CD-RW drive
  • A motherboard
  • A hard drive

PC Ports

Standard PC

Page 40 of the Philips Patient Monitor (M2540) user manual — VCR-Option Components in the PC
Page 40 · click to enlarge
Page 40
Theory of OperationTR Boards

The VCR option includes several cards inside the PC and an adapter that connects to the SVGA

  • An SVGA-to-S-Video card that converts the SVGA signal from the graphics adapter board (in

communicates with the ultrasound system through a USB cable to the PC motherboard. See

  • A magneto-optical disk (MOD) drive that allows archiving of images, studies, and reports on
  • An SVGA adapter that connects the graphics adapter board to the SVGA-to-TV video card
  • The Physio module that provides ECG monitoring capability. The drive bay’s power connec-

tor powers the module, but the module does not use the drive bay’s data port. The module

“Physio Module Option” on page 43, and Figure 4-3 on page 51 for detailed information on

TR Boards The M2540A’s e-box contains two transmit-and-receive (TR) boards. Each has a connector that

plugs into the motherboard and another connector that passes through the motherboard and

representation of how the VCR option’s components function in the ultrasound system.

See “VCR Option Parts” on page 198 for a list of all the components of the VCR option.

output of the PC’s graphics adapter board. See Figure 4-1 on page 48 for a detailed

  • A sound card that receives audio from the VCR, and supplies sound to the VCR

The following cards are part of the VCR option, and reside in the PC’s PCI slots:

removable disks. The MOD uses the drive bay’s data and power ports.

  • A video capture board that receives S-Video signals from the VCR

The PC’s drive bays accept the following components:

VCR-Option Components in the PC

Optional Drive-Bay Components

the AGP slot) to S-Video for the VCR

and to the monitor

the Physio module.

Page 41 of the Philips Patient Monitor (M2540) user manual — VCR-Option Components in the PC
Page 41 · click to enlarge
Page 41
Theory of OperationTR Boards

input bus. The last TR board in this chain outputs the final summed radio-frequency (RF) data to

are daisy-chained together to sum the outputs. The RF data output of one TR board is the input

to the next TR board, and the outputs of the TR boards are summed to form the received beam.

and times each of the analog drivers. The delay generators receive the appropriate delay values

The received signal is preamplified, TGC amplified, low-pass filtered, and digitized (in all modes

After being digitized, the signal on each line is delayed appropriately. The outputs of all channels

filters, 10-bit A/D converters, and one eighth of a digital beam-forming circuit. Each TR board

The BPAP board generates the timing to start and stop each firing line, and sends the transmit

transducer. Analog drivers generate the transmitted RF data. A digital timing generator delays

except CW). The BPAP board generates TGC control signals through the demodulator board.

plugs into the distribution board. The two boards form electronic beams of the transmit and

sends on its output bus a value that is the sum of the data from its 32 channels plus its 18-bit

The transmit and receive beam-forming coefficients for acoustic lines are downloaded on the

receive signals. Each TR board contains 64 transmitters and 32 preamps multiplexed into 64

receive channels. This allows multiplexing the system’s 64 active channels into 128 element

arrays. Each receive channel consists of time gain compensation (TGC) amplifiers, low pass

The CW Doppler signal is mixed to an analog intermediate frequency (IF) signal. The CW

signal to the TR board, through the distribution board and the connector modules to the

from the BPAP for each channel. The system power supply controls the amplitude of the

Doppler output is a single differential signal routed to the demodulator board.

The summed output signal (RF DATA) goes to the demodulator board.

coefficient bus from the BPAP board.

the demodulator board.

transmitted waveform.

Page 42 of the Philips Patient Monitor (M2540) user manual — Omni TEE
Page 42 · click to enlarge
Page 42
Theory of OperationBPAP Board

motor. It also measures the position of the angle-position sensor. The signal is digitized and sent

  • Houses electronic switching that allows selection of a single active transducer from the four

BPAP Board The beam processor/acoustic processor (BPAP) board contains two processors. One of them

Circuitry The Omni Probe control circuitry on the BPAP board supplies power to the TEE transducer's

  • Houses the transesophageal echo cardiography (TEE) transducer interface circuitry (TEE
  • Receives control data sets from the PC, processes them for specific register values, and

The beam processor circuitry monitors thermistors in the transducer and insures that high

controls beam-forming, line-to-line operation. The other handles the processing of scan

voltage is not applied to the transmitter unless the thermistor values are within limits.

  • Provides the interface between the PC and the demodulator board
  • Controls the high-voltage output from the system power supply
  • Provides the interface between the PC and the TR boards
  • Detects the presence of standalone transducers

motor control and temperature monitoring)

to the APIO board for monitoring by the PC.

conversion, color flow, and Doppler signals.

  • Houses the transducer ID circuitry

The BPAP board does the following:

sends them to the TR boards

that can be connected

Omni TEE

Page 43 of the Philips Patient Monitor (M2540) user manual — Physio Module
Page 43 · click to enlarge
Page 43
Theory of OperationDemodulator Board

and CW phased and non-imaging Doppler modes, and color flow mode. It sends a baseband data

  • The independent transducer CW Doppler electronics and steerable CW detector functions

and the e-box. The APIO board also handles input from the foot switch’s three control buttons

The demodulator board performs image and Doppler demodulation for BMode, MMode, PW

APIO Board The acoustic processor I/O (APIO) board is the token ring interface between the system’s PC

(via the I/O panel) and forwards them to the PC. The board also generates trigger signals for

The self-contained Physio module resides in one of the PC’s drive bays. The module receives

power from the drive bay’s power connector, but it communicates with the system through a

electromagnetic interference. See Figure 4-3 on page 51 for a detailed representation of the

  • The 120 MHz system clock. The clock signal is distributed to the TR and BPAP boards.

USB cable to the PC’s motherboard. A shielded and grounded enclosure prevents

  • Digital filters and mixers for demodulating 2D, color flow, and PW RF data
  • The R-wave pulse from the physio module that it sends to the BPAP board
  • Acoustic data from the BPAP board for the acoustic imaging display
  • The DC on-off bit that it sends to the power supply
  • Commands that program the BPAP board at startup

The demodulator board contains the following:

The board handles these signals from the PC:

  • Circuitry that generates TGC signals

module’s functions and connections.

stream to the BPAP board.

legacy printers.

Physio Module

Demodulator

Option

Board

Page 44 of the Philips Patient Monitor (M2540) user manual — Primary
Page 44 · click to enlarge
Page 44
Theory of OperationSystem Motherboard
  • Process electrocardiogram (ECG) signals from patient electrodes and deliver the data to the

This board connects all data paths between the TR boards, the BPAP board, and the distribution

The distribution board passes transducer signal inputs to the imaging system. The system power

  • TEE motor-control and temperature-sense signals from the BPAP to and from the TEE trans-
  • The ECG-input connector is a round, 12-pin connector that is isolated from chassis ground.

board. It is also the distribution point for DC power and ground to all boards (the BPAP board

and distribution board also receive power directly from the system power supply.). It also sends

  • Send a detected R-wave trigger to the APIO board for frame triggering in various scanning
  • A ¼-inch phone plug jack accepts auxiliary input and has both signal and shield grounds.
  • Transducer identification signals from the BPAP board to and from the transducer

supply provides high voltage directly to the distribution board to fire transducers.

  • Transmit and receive signals from the TR boards to the active transducer

Connectors The Physio module has two connectors on its front panel:

Functions The module’s primary functions are as follows:

  • Provide gain control for ECG waveforms

Signals that pass through this board include

filtered high voltage to the TR boards.

Motherboard

Distribution

Connector

Board and

Modules

Primary

System

M2540A’s PC · modes · ducer

Page 45 of the Philips Patient Monitor (M2540) user manual — Module Types
Page 45 · click to enlarge
Page 45
Theory of OperationI/O Panel

rotary and slide controls, and the trackball. A USB cable carries control inputs from the control

I/O Panel The I/O panel bears four connectors and two ground points. The main system ground used for

The system control panel contains the circuitry to process control inputs, the system keyboard,

entire assembly, including circuit boards, bezel assembly, and labels, is available as a replaceable

panel to the PC. An auxiliary cable carries power from the system power supply. The system

The four transducer module bays accept the following types of modules in any combination:

Circuit boards inside the system control panel are not field serviceable or replaceable. The

All control inputs and LEDs are controlled by the system’s PC, and their functions are

control panel circuit board also carries the amplifiers for the system’s stereo speakers.

safety testing, and an equipotential post are the two ground points on the panel.

  • A BNC connector that carries composite video signals for a legacy printer

See Table 13-2 on page 179 for a list of the modules and their part numbers.

  • A BNC connector that carries trigger signals for a legacy printer
  • An RJ-45 (CAT 5) connector for connection to a LAN

determined by resident control-panel and system software.

  • A 9-pin D connector for the optional foot switch
  • Pencil-type transducer connector

The I/O panel’s connectors are

  • Advance Vision connector
  • Cannon connector
  • Blank module

Control Panel

Module Types

Transducer

Connector

System

Page 46 of the Philips Patient Monitor (M2540) user manual — Power Supply
Page 46 · click to enlarge
Page 46
Theory of OperationPower Supply

Power Supply The M2540A power supply is a self-contained unit that accepts AC input from 90 to 264 V, at 50

also provides all DC power required by system components. No switch settings are required to

configure the power supply for different AC mains; the power supply automatically adjusts to use

the power connected to it. Except for the 120 VAC to the PC, all power output is activated by a

signal from the PC. The system cooling fan is the only serviceable part on the power supply. See

Figure 4-1 on page 48 for details of the power supply’s outputs. The power supply activates when

it receives an “AWAKE_N” signal, which is generated by the PC, and then sent through the BPAP

  • Provides DC power to all system electronics through the system motherboard and the BPAP
  • Provides three isolated 120 VAC switched outlets to power the monitor and two peripherals

part. Rotary and sliding knobs, rotary encoders for the knobs, the trackball assembly, and the

or 60 Hz, and supplies 120 VAC at 60 Hz to the system’s PC, monitor, and two peripherals. It

  • Provides high voltage to the distribution board for multiplexing transducers
  • Provides an isolated 120 VAC unswitched outlet to power the internal PC
  • Isolates AC mains from secondary AC and DC circuitry
  • Provides DC power directly to the system cooling fan
  • Transforms primary voltages to secondary voltages

trackball cable are all available as separate parts.

The system power supply does the following:

board to the power supply.

  • Filters line AC

board

Page 47 of the Philips Patient Monitor (M2540) user manual — Functional Block
Page 47 · click to enlarge
Page 47
Theory of OperationFunctional Block Diagrams
  • Figure 4-1 on page 48 shows the ultrasound system with the VCR option installed. This dia-
  • Figure 4-2 shows the e-box and the system power supply and their connections to other

gram does not include the e-box or the system’s power supply. Those components are

  • Figure 4-3 on page 51 shows the optional Physio module and its internal connections.

This section contains block diagrams of the system’s components and their connections.

Figure 4-1 shows the M2540A system with the VCR option installed.

The section includes the following block diagrams:

components of the ultrasound system.

shown in Figure 4-2 on page 49.

Functional Block

Functional

Diagrams

Diagram

System

Block

Page 48 of the Philips Patient Monitor (M2540) user manual — Functional Block Diagrams
Page 48 · click to enlarge
Page 48
Theory of OperationFunctional Block Diagrams
Philips Patient Monitor (M2540) — Functional Block Diagrams diagram (page 48)
Fig. 7 · Functional Block Diagrams

Figure 4-1 System Block Diagram with VCR Option

USB printer · To remote

Page 49 of the Philips Patient Monitor (M2540) user manual — e-box and Power
Page 49 · click to enlarge
Page 49
Theory of OperationFunctional Block Diagrams

Figure 4-2 shows the system’s power supply and the e-box and their connections to each other

Philips Patient Monitor (M2540) — e-box and Power Supply Functional Block Diagram diagram (page 49)
Fig. 8 · e-box and Power Supply Functional Block Diagram

To APIO board in PC

Figure 4-2 e-box and Power Supply Block Diagram

and other components.

e-box and Power

Functional Block

Diagram

Supply

To system

Page 50 of the Philips Patient Monitor (M2540) user manual — Token Ring Cable
Page 50 · click to enlarge
Page 50
Theory of OperationFunctional Block Diagrams

The token ring cable carries the following signals between the BPAP board and the APIO board:

  • On and off signal that switches on the outputs from the power supply (the 120 VAC output
  • Program enable and program done (on separate circuits)
  • Ring in and ring out (on separate circuits)
  • R-wave communication

for the PC is always on)

Token Ring Cable

Page 51 of the Philips Patient Monitor (M2540) user manual — Functional Block
Page 51 · click to enlarge
Page 51

Figure 4-3 shows how the Physio module functions, its internal connections, and connections to

Theory of OperationFunctional Block Diagrams

A/D

EZ-USB

M

U

USB

0.35 to 30Hz

Figure 4-3 Physio Module Block Diagram

other components.

Functional Block

Diagram

Physio

Show 42 more labels / short text on this page
nlddac1 · dacdata · r-raw · R-wave · detector · converter · nadccs · spi-ck · auxsel spi-di · phy1 · phy1 · phy2 · spi-do · Phy1 offset · gain adjust · Phy2 offset · gain adjust · dacdata · controller · and · and · nlddac2 · To PC motherboard · Bandpass · R-wave trig · filter · AUX In · clk24 · islout X216 · Clock · divider · To APIO · Signal · isolation · Power · isolation · Isolated front end · Differential · amp · driver · and · ECG leads
Page 52 of the Philips Patient Monitor (M2540) user manual — Safety Testing:
Page 52 · click to enlarge
Page 52
Theory of OperationTransducer Safety Testing: Test Setup and Theory

This section provides background and supporting information for the transducer leakage tests in

The test for transthoracic and endocavity transducers differs from the test for TEE transducers

insulation and provides an electrical path between the submerged lead wire and the grounded

In transducer safety tests, a container filled with saline solution functions as a conductive

medium (see Figure 4-4 on page 53). The solution penetrates any faults in the transducer

Section 6, “Performance Tests”; this information is not required to perform the tests.

only by the extent that the transducers are submerged in the test solution.

inner transducer shield.

Safety Testing:

Transducer

and Theory

Test Setup

Page 53 of the Philips Patient Monitor (M2540) user manual — 7ASW030-1
Page 53 · click to enlarge
Page 53
Theory of OperationTransducer Safety Testing: Test Setup and Theory

Closed for 2nd single fault condition

7ASW030-1

Philips Patient Monitor (M2540) — 7ASW030-1 diagram (page 53)
Fig. 9

Open for 1st single fault condition

Neutral conditionClosed for normal condition

Case Leakage, Ground Conductor for ground wire leakage test Mode selector on safety analyzer:

Figure 4-4 Transducer Leakage Current Test Setup and Theory Diagram

Ground condition:

ECG for transducer leakage test

Closed for 1st single fault condition

Open for 2nd single fault condition

Closed for normal condition

Show 4 more labels / short text on this page
Neutral · Hot · See Key for · Figure 4-4
Page 54 of the Philips Patient Monitor (M2540) user manual — Z =
Page 54 · click to enlarge
Page 54
Theory of OperationTransducer Safety Testing: Test Setup and Theory
  • For TEE transducers, submerge all of the flexible shaft that would normally enter the body of

Leakage current Ichassis, driven by the line supply, flows through the stray capacitance C between

Analyzer. Select ECG for transducer leakage and Case Leakage, or select

Impedance between the transducer’s metal parts and the test electrode:

  • For transthoracic and endocavity transducers, submerge the head and 5 cm of the cable,

through the ECG lead wire. (This is the mode selector on the Safety

S = Switch that connects the ammeter either directly to the chassis or

C = Stray capacitance from the system's power wiring to the chassis.

≈ 800 KΩ if the outer insulating layer is intact

< 500 Ω if the outer layer is compromised

Theory of the Transducer Leakage Current Test

Ground Conductor for chassis leakage.)

the patient: 100 cm for adult TEE, 60 cm for pediatric TEE.

A = Microammeter in the safety analyzer

being careful to not submerge the connector.

the primary wiring and the system’s metal chassis.

Table 4-1 Key for Figure 4-4

Imeasured = Leakage current

Z =

Insertion depth:

Page 55 of the Philips Patient Monitor (M2540) user manual — 7ASW030-2
Page 55 · click to enlarge
Page 55
Theory of OperationTransducer Safety Testing: Test Setup and Theory

The leakage current normally flows from the chassis safely out through the ground wire. If there

is a fault in the transducer insulation, some of the current follows this path and is measured by

the safety analyzer. This Itransducer is still fairly low, unless the chassis is not properly grounded.

Figure 4-5 Transducer Leakage Current Test Diagram for the Normal Condition

Philips Patient Monitor (M2540) — Theory of the Transducer Leakage Current Test diagram (page 55)
Fig. 10 · Theory of the Transducer Leakage Current Test

7ASW030-2

Show 4 more labels / short text on this page
system · Ground · Neutral · Hot
Page 56 of the Philips Patient Monitor (M2540) user manual — 7ASW030-4
Page 56 · click to enlarge
Page 56
Theory of OperationTransducer Safety Testing: Test Setup and Theory

When an open ground condition is imposed, all Ichassis is forced through the transducer. This

Figure 4-6 Transducer Leakage Current Test Diagram for 1st Single Fault Condition

Figure 4-7 Transducer Leakage Current Test Diagram for 2nd Single Fault Condition

Philips Patient Monitor (M2540) — Theory of the Transducer Leakage Current Test diagram (page 56)
Fig. 11 · Theory of the Transducer Leakage Current Test

7ASW030-3

Philips Patient Monitor (M2540) — 7ASW030-3 diagram (page 56)
Fig. 12

7ASW030-4

Itransducer is still fairly low unless its sheath is compromised.

Show 8 more labels / short text on this page
system · system · Ground · Neutral · Ground · Neutral · Hot · Hot
Page 57 of the Philips Patient Monitor (M2540) user manual — 7ASW030-5
Page 57 · click to enlarge
Page 57
Theory of OperationTransducer Safety Testing: Test Setup and Theory
  • If the transducer sheath is intact, its resistance is high, and I measured in the transducer leakage

flows through the transducer. This Itransducer is still fairly low, unless the chassis is not properly

In the 1st single fault condition transducer leakage test (Figure 4-6 on page 56), all the leakage

current Ichassis flows through the analyzer by way of the transducer. In that condition, only the

higher potential for leakage current Ichassis. Most flows safely through the ground wire; some

With the transducer circuit disconnected, all of the leakage current Ichassis flows through the

Imposing an open neutral condition prevents all system current from flowing. This creates a

Figure 4-8 Ground Wire Leakage Test Diagram (for Comparison with 1st Single Fault

test is low when compared with Imeasured in the ground wire leakage test.

Philips Patient Monitor (M2540) — Theory of the Transducer Leakage Current Test diagram (page 57)
Fig. 13 · Theory of the Transducer Leakage Current Test

7ASW030-5

resistance or impedance of the respective paths varies.

analyzer by way of the ground wire.

grounded.

Show 5 more labels / short text on this page
system · Condition) · Neutral · Ground · Hot
Page 58 of the Philips Patient Monitor (M2540) user manual — Theory of the Transducer Leakage Current Test
Page 58 · click to enlarge
Page 58
Theory of OperationTransducer Safety Testing: Test Setup and Theory

Transducer leakage current equaling 80% or more of ground wire leakage current indicates a

causing significant Itransducer, and yet that leakage current could still be within acceptable limits.

  • If the transducer sheath is compromised, its resistance is close to that of the ground wire.

Sheath integrity is tested this way because there could be a break in the transducer sheath

Comparison to the ground wire leakage current is the only way to ensure sheath integrity.

fault or break in the transducer insulation.

Page 59 of the Philips Patient Monitor (M2540) user manual — Power Cord and
Page 59 · click to enlarge
Page 59
InstallationOverview
3. Rotate the twivel base to its proper position. (The monitor cables pass through the rear of the twivel base.)
2. Loosen the screw that secures the cord retainer , rotate the retainer 180 degrees, and tighten the screw.
  • When two peripherals are installed, the smaller one must be on top of the larger one.
1. Attach the system power cord to the receptacle at the bottom rear of the system.
5. Connect the VGA cable the and monitor power cable to the monitor.

Overview This chapter includes procedures for installing and setting up the M2540A.

Complete the following steps to assemble the M2540A ultrasound system:

  • No more than two peripherals can be mounted on the cart.

Installation When installing peripherals, keep the following points in mind:

➤Installing the Power Cord and Monitor

The following sections are in this chapter:

  • “Assemble the System” on page 59
  • “System Configuration” on page 64
4. Install the monitor on the twivel.
  • “Customer Training” on page 88
  • “System Startup” on page 62

5 Installation

Power Cord and

Assemble the

Installation

Peripherals

System

Monitor

Page 60 of the Philips Patient Monitor (M2540) user manual — Installing a VCR
Page 60 · click to enlarge
Page 60
InstallationAssemble the System
4. If another peripheral is part of the installation, place the VCR peripheral tray on the top of the VCR.
2. Connect the power cable , audio input cable, audio output cable , and S-video cable to the VCR.
  • Peripherals should mount toward the rear of the system, with about one centimeter
8. Save the VCR manual and any other documentation from the shipping box.

between the lower peripheral and the inside edge of cart top.

3. Connect the serial cable to the VCR, and screw it in place.
1. Loosen the peripheral strap , and slide the VCR under it.
5. Tighten the peripheral strap on top of the peripherals.
7. Set the VCR switches to the following settings:
  • Audio Out right switch: Norm
  • Audio Out left switch: Mix

➤Installing a Sony Color Printer

6. Turn on the VCR power switch .
  • Mode Lock: Off
  • Input: S-Video
  • S-VHS: Auto

➤Installing a VCR

  • Menu: Off
Page 61 of the Philips Patient Monitor (M2540) user manual — Installing a Sony Black-and-White Printer
Page 61 · click to enlarge
Page 61
InstallationAssemble the System
3. Take a cable stop bracket from the accessories shipping box and slip it onto the peripheral strap.
4. Hook the bracket in a slot of the cart top or the peripheral tray so the bracket is as close as possible to the printer.
3. Take a cable stop bracket from the accessories shipping box and slip it onto the peripheral strap.
1. Loosen the peripheral strap , and slide the printer under it.
1. Loosen the peripheral strap , and slide the printer under it.
2. Connect the power cable and a USB cable to the printer.
7. Save the following material from the printer’s shipping box:
2. Connect the power cable and a USB cable to the printer.

➤Installing a Sony Black-and-White Printer

6. Switch the printer power switch on .
  • Any other documentation

black-and-white printer.

  • Cleaning ribbon
  • Printer manual
  • Printer paper
  • Print ribbon
5. Tighten the strap .
Page 62 of the Philips Patient Monitor (M2540) user manual — Monitor Signal
Page 62 · click to enlarge
Page 62
InstallationSystem Startup
4. Hook the bracket in a slot of the cart top or the peripheral tray so the bracket is as close as possible to the printer.

Voltage When shipped from the factory, the monitor signal input voltage is set to 1.0 V. This must be

2. Press the two right-hand monitor buttons , and hold them in while reconnecting the power cord.
6. On the back of the printer, set dip switch number 5 to the UP position. All the other switches should be down.

Complete the procedures in this section to set up the system for the customer’s use.

8. Save the following material from the printer’s shipping box:
1. Disconnect the power cord from the rear of the monitor.
2. Press the power switch on the front of the system’s PC.
1. Connect the system power cord to a wall outlet.

➤Setting the Monitor Signal Input Voltage

reset to 0.7 V, using the following procedure:

7. Switch the printer power switch on.
  • Any other documentation

➤Powering On the System

  • Thermal paper
  • Cleaning sheet
  • Printer manual
5. Tighten the strap.

Monitor Signal

System

Startup

Page 63 of the Philips Patient Monitor (M2540) user manual — System Startup
Page 63 · click to enlarge
Page 63
InstallationSystem Startup
4. The two right-hand monitor buttons scroll up or down in the OSD Main Menu. Use them to select Advanced Controls .
8. Use the two right-hand monitor buttons to select SAVE & RETURN .
6. Use the two right-hand monitor buttons to select INPUT 0.7V .
3. Press the two center monitor buttons .
Philips Patient Monitor (M2540) — System Startup diagram (page 63)
Fig. 14 · System Startup
Philips Patient Monitor (M2540) — System Startup diagram (page 63)
Fig. 15 · System Startup

The Advanced Controls menu displays.

5. Press the far left monitor button .
7. Press the far left monitor button .

The OSD Main Menu displays.

Figure 5-2 Advanced Controls Menu

Figure 5-1 OSD Main Menu

Page 64 of the Philips Patient Monitor (M2540) user manual — Printer
Page 64 · click to enlarge
Page 64
InstallationSystem Configuration
2. Set the paper type switch inside the printer’s paper drawer to the loaded paper type. (The paper type is marked on the paper’s foil wrapper, and on the tube at the center of the paper
10. On the OSD Main menu, use the two right-hand monitor buttons to select EXIT OSD .
3. Turn the printer power on, to move the print head from its shipping park position.

Administrator must be available to supply configuration data.

The following procedures configure the system’s software:

1. Load thermal paper in the printer’s paper drawer.

Preparation ➤Preparing the Sony Black-and-White Printer

11. Press the far left monitor button to exit OSD.

➤Preparing the Sony Color Printer

1. Load paper in the printer’s paper tray.
9. Press the far left monitor button .
2. Insert the paper tray in the printer.
4. Load the print ribbon .

roll.)

Configuration

System

Printer

Page 65 of the Philips Patient Monitor (M2540) user manual — Localization
Page 65 · click to enlarge
Page 65
InstallationSystem Configuration
Philips Patient Monitor (M2540) — Localization diagram (page 65)
Fig. 16 · Localization
3. Click the Top Border button. The Top Border window opens.

Localization ➤Setting the Localization Parameters

1. Press Setup . The Setup window displays.
2. Click the System tab.

Figure 5-3 Setup Window

Page 66 of the Philips Patient Monitor (M2540) user manual — Localization
Page 66 · click to enlarge
Page 66
InstallationSystem Configuration
5. On the Setup window, click the Date/Time button.
Philips Patient Monitor (M2540) — Localization diagram (page 66)
Fig. 17 · Localization

The Date and Time Properties window opens.

4. Enter the institution name, and click OK .

Figure 5-4 Top Border Window

Page 67 of the Philips Patient Monitor (M2540) user manual — Localization
Page 67 · click to enlarge
Page 67
InstallationSystem Configuration
10. On the Setup window, click the Locale button. The Regional and Languages Options window opens.
9. Change the settings to the local date and time, and click OK .
7. Choose the local time zone from the drop-down list.
Philips Patient Monitor (M2540) — Localization diagram (page 67)
Fig. 18 · Localization

Figure 5-5 Date and Time Properties Window

8. Click the Date & Time tab.
6. Click the Time Zone tab.
Page 68 of the Philips Patient Monitor (M2540) user manual — NOTES
Page 68 · click to enlarge
Page 68
InstallationSystem Configuration
12. If the number, currency, time, or date formats in the selected configuration set do not match the site locale, click Customize and change the incorrect settings.
  • Customizing the date format to one that’s longer than any in the drop-down list may cause
11. Click the drop-down list in the Standards and Formats section, and select the configuration set in the list that most nearly matches the site locale.

Figure 5-6 Regional and Languages Options Window (Regional Options)

Philips Patient Monitor (M2540) — Localization diagram (page 68)
Fig. 19 · Localization

dates in the display to be truncated.

NOTES

Page 69 of the Philips Patient Monitor (M2540) user manual — Localization
Page 69 · click to enlarge
Page 69
InstallationSystem Configuration
  • The locale selected in Standards and formats section ( step 11 ) determines what language the

online Help displays in, if that language is available. If it’s not, Help uses the same language as

13. Click the Location drop-down list, and select the location that most nearly matches the site.
15. Click the Language used in menus and dialogs menu, and select the local language.
14. On the Regional and Languages Options window, click the Languages tab.

Figure 5-7 Regional and Languages Options Window (Languages)

Philips Patient Monitor (M2540) — Localization diagram (page 69)
Fig. 20 · Localization

the rest of the system.

Page 70 of the Philips Patient Monitor (M2540) user manual — Localization
Page 70 · click to enlarge
Page 70
InstallationSystem Configuration

system. You may be prompted to restart or to “use existing files.” Click Yes in either

17. Click the drop-down menu and select the language for non-Unicode programs . The selection should be the one that most closely matches that chosen in step 13 .
18. Click OK to close the Regional and Languages Options window.

Figure 5-8 Regional and Languages Options Window (Advanced)

Philips Patient Monitor (M2540) — Localization diagram (page 70)
Fig. 21 · Localization
16. Click the Advanced tab.

case.

Page 71 of the Philips Patient Monitor (M2540) user manual — Options
Page 71 · click to enlarge
Page 71
InstallationSystem Configuration

Installation The M2540A leaves the factory with all purchased options enabled. If an option must be enabled

Philips Patient Monitor (M2540) — Options Installation diagram (page 71)
Fig. 22 · Options Installation
2. Click the Options... button. An installation window opens.
1. On the Setup window, click the Options tab.

in the field, complete the following procedure:

Figure 5-9 Setup Window (Options)

Options

➤Installing Options

Page 72 of the Philips Patient Monitor (M2540) user manual — Options Installation
Page 72 · click to enlarge
Page 72
InstallationSystem Configuration
3. Select an application to enable by clicking it so it is highlighted.
6. Repeat steps 3 through 5 to enable any additional options.
Philips Patient Monitor (M2540) — Options Installation diagram (page 72)
Fig. 23 · Options Installation
7. When all the required options are installed, click OK .
5. Enter the access code in the window, and click OK .

The selected option is enabled.

Figure 5-10 Options Installation Window

4. Click Install .
Page 73 of the Philips Patient Monitor (M2540) user manual — Configuration
Page 73 · click to enlarge
Page 73
InstallationSystem Configuration

Complete the following procedures to configure the system for DICOM communications over a

network. The procedure requires information from the customer’s network administrator.

Philips Patient Monitor (M2540) — DICOM Network Configuration diagram (page 73)
Fig. 24 · DICOM Network Configuration

➤Changing the System’s Network Parameters

1. Press the Setup key. The Setup window displays.

The DICOM Setup window displays.

2. Click the DICOM button.

Figure 5-11 DICOM Setup Window

Configuration

Network

DICOM

Page 74 of the Philips Patient Monitor (M2540) user manual — Using the Field Service Options Window for Advanced Network Settings
Page 74 · click to enlarge
Page 74
InstallationSystem Configuration

➤Using the Field Service Options Window for Advanced Network Settings

following procedure to change the network settings.

1. On the Setup window, click the Options tab.
5. Enter the settings provided by the customer.
3. Click the This System tab.
4. Click Network Settings .
6. Click OK .
Page 75 of the Philips Patient Monitor (M2540) user manual — Using the Field Service Options Window for Advanced Network Settings
Page 75 · click to enlarge
Page 75
InstallationSystem Configuration

The Field Service Options window opens.

Figure 5-12 Setup Window (Options)

2. Click Service .
Page 76 of the Philips Patient Monitor (M2540) user manual — Using the Field Service Options Window for Advanced Network Settings
Page 76 · click to enlarge
Page 76
InstallationSystem Configuration

work as an ultrasound imaging device. This is true even if no settings are changed.

The display changes to Windows Classic format.

Philips Patient Monitor (M2540) — Using the Field Service Options Window for Advanced Network Settings diagram (page 76)
Fig. 25 · Using the Field Service Options Window for Advanced…
10. Enter the values supplied by the customer.
6. Double-click Local Area Connections .
5. Double-click Network Connections .
8. Click Internet Protocol (TCP/IP) .

The Control Panel window opens.

Figure 5-13 Field Service Options Window

4. Click Switch to Classic View .
3. Click Control Panel .
7. Click the General tab.
9. Click Properties .
Page 77 of the Philips Patient Monitor (M2540) user manual — Adding Server Information
Page 77 · click to enlarge
Page 77
InstallationSystem Configuration
3. Enter the server information supplied by the network administrator in the appropriate fields.
Philips Patient Monitor (M2540) — Adding Server Information diagram (page 77)
Fig. 26 · Adding Server Information
1. On the DICOM Setup window, click the Servers & Roles tab.

Figure 5-14 DICOM Setup Window (Servers & Roles)

2. In the Servers section, click New .

➤Adding Server Information

Page 78 of the Philips Patient Monitor (M2540) user manual — NOTES
Page 78 · click to enlarge
Page 78
InstallationSystem Configuration
  • The value in the Name: field need not be the server’s DNS or network name. It is used only
  • The value in the Host: field is the server’s IP address, or the name that the DNS server

The M2540A tests communication with the server, and opens a window that shows the

1. In the Roles section under the Servers & Roles tab, click Modify .
5. Make sure the network cable is connected, and click Ping .

to identify the server in error messages and dialog boxes.

➤Assigning Servers to DICOM Roles

resolves to an IP address.

4. Click Done .

result.

NOTES

Page 79 of the Philips Patient Monitor (M2540) user manual — Configuring Autodelete
Page 79 · click to enlarge
Page 79
InstallationSystem Configuration
3. To modify or verify attributes of storage or print servers, click Advanced next to the server’s drop-down list.

The following procedure configures how the system automatically deletes studies.

Philips Patient Monitor (M2540) — Configuring Autodelete diagram (page 79)
Fig. 27 · Configuring Autodelete
2. Select the correct server for each role from the drop-down lists.
1. On the DICOM Setup window, click the This System tab.

Figure 5-15 DICOM Setup Window (Servers & Roles)

4. Click Done .

➤Configuring Autodelete

Page 80 of the Philips Patient Monitor (M2540) user manual — Configuring Autodelete
Page 80 · click to enlarge
Page 80
InstallationSystem Configuration
Philips Patient Monitor (M2540) — Configuring Autodelete diagram (page 80)
Fig. 28 · Configuring Autodelete
2. Click Disk Full Strategy The Disk Full Strategy window opens.

Figure 5-16 DICOM Setup (This System)

Page 81 of the Philips Patient Monitor (M2540) user manual — Configuring Logging
Page 81 · click to enlarge
Page 81
InstallationSystem Configuration
1. On the DICOM Setup window, click the Diagnostics tab.
Philips Patient Monitor (M2540) — Configuring Logging diagram (page 81)
Fig. 29 · Configuring Logging
3. Set the parameters to the customer’s preference.
4. Click OK to close the window.

Figure 5-17 Disk Full Strategy Window

2. Click Logging .

➤Configuring Logging

Page 82 of the Philips Patient Monitor (M2540) user manual — Configuring Logging
Page 82 · click to enlarge
Page 82
InstallationSystem Configuration
5. Under System Log, make sure the Logging level is set to Error Level Logging .
3. Under DICOM log contents, select No logging (disables logging) .
Philips Patient Monitor (M2540) — Configuring Logging diagram (page 82)
Fig. 30 · Configuring Logging
4. Under Service log, select Enable Service logging .
6. Click OK to close the DICOM Setup window.

Figure 5-18 DICOM Setup Window (Diagnostics)

Page 83 of the Philips Patient Monitor (M2540) user manual — Configuration
Page 83 · click to enlarge
Page 83
InstallationSystem Configuration

The following procedures configure the system for the installed peripherals and system

2. Select a peripheral device from the Record, Rec2, and Rec3 drop-down lists.
Philips Patient Monitor (M2540) — System Hardware Configuration diagram (page 83)
Fig. 31 · System Hardware Configuration
1. On the Setup window, click the Peripherals tab.

➤Assigning Record Buttons

Figure 5-19 Setup Window (Peripherals)

hardware.

Configuration

Hardware

System

Page 84 of the Philips Patient Monitor (M2540) user manual — Assigning Option Buttons
Page 84 · click to enlarge
Page 84
InstallationSystem Configuration
1. On the Setup window, click the Options tab.

The Define Option Keys window opens.

➤Assigning Option Buttons

Figure 5-20 Setup window (Options)

2. Click Keyboard
Page 85 of the Philips Patient Monitor (M2540) user manual — VCR configuration
Page 85 · click to enlarge
Page 85
InstallationSystem Configuration

The system is configured at the factory for NTSC format. To reconfigure the system for PAL

3. Click the drop-down list for each option key, and select the appropriate functions.
Philips Patient Monitor (M2540) — VCR configuration diagram (page 85)
Fig. 32 · VCR configuration
1. On the Setup window, click the Peripherals tab

➤Configuring the System for PAL Format

format, complete the following procedure:

Figure 5-21 Define Option Keys Window

VCR configuration

features.

4. Click OK .
Page 86 of the Philips Patient Monitor (M2540) user manual — VCR configuration
Page 86 · click to enlarge
Page 86
InstallationSystem Configuration

The Install Software Drivers window opens.

2. Click Install Software Drivers .

Figure 5-22 Setup Window (Peripherals)

3. Click Video Installation .
5. Select PAL .
4. Click OK .
6. Click OK .
Page 87 of the Philips Patient Monitor (M2540) user manual — Configuring for printers
Page 87 · click to enlarge
Page 87
InstallationSystem Configuration

Cardiologists prefer a bluer tint to the monitor display than radiologists do. Pressing the two left

3. To change the setting, press the two left buttons again, while the current setting is displayed.

The drivers for Sony UP-D895, Sony UP-D21MD, HP920, and HP1200 printers are loaded

The driver is reinstalled with the correct configuration. This takes about 20 seconds.

1. In normal ultrasound mode, press the two left monitor buttons at the same time.

automatically. These printers are ready to use when connected to a USB port.

monitor buttons together toggles between the two color settings.

2. A window displays the current color temperature setting.

The new setting takes effect after the system is restarted.

  • Setting 2 is appropriate for non-cardiology use.

➤Setting the Monitor Color Temperature

  • Setting 1 is appropriate for cardiology use.
7. Click Close to close the Setup window.

To set the monitor color temperature:

Configuring for printers

Page 88 of the Philips Patient Monitor (M2540) user manual — Customer
Page 88 · click to enlarge
Page 88
InstallationCustomer Training
  • Explain the importance of letting the system shut down completely before disconnecting the
  • Demonstrate how to clean the air filter ( “Cleaning the Air Filter” on page 131 ), and explain
  • Instruct the customer how to back up and restore presets (see “Backing Up and Restoring

Presets” on page 143). Make a backup floppy disk of the customer’s presets and configura-

tions, and store it behind the lower enclosure’s left rear door. If the software is reloaded

  • If the system is connected to a DICOM network, show the customer the job queue, and
  • Demonstrate how to remove and clean the trackball. (See “Cleaning the Trackball” on

The customer needs some instruction in maintenance and use of the system:

later, the backup floppy disk lets you quickly restore settings and options.

explain that the queue is empty if all studies have transferred successfully.

the importance of the filter to system performance.

power cord.

page 132.)

Customer

Training

Page 89 of the Philips Patient Monitor (M2540) user manual — NOTES
Page 89 · click to enlarge
Page 89
Performance TestsOverview

Overview This chapter includes all test procedures for diagnosing problems with the M2540A and assuring

The tests described on the following pages are part of a comprehensive preventive maintenance

  • The system power supply is a self-contained unit that is safety-tested by its manufacturer. No
  • After any service or maintenance, perform the required tests to assure that system safety
  • The Chassis to Ground Resistance test checks the entire system for electrical resistance
  • The Ground Wire Leakage Current test checks the entire system for leakage on ground

between the chassis and ground. See “Chassis to Ground Resistance Test” on page 90.

program. They detect abnormalities that could prove dangerous to patient or operator.

further safety testing is needed when the power supply is installed or replaced.

leads. See “Ground Wire Leakage Current Test” on page 92.

  • Inspect power cords frequently, especially at their ends.

The following safety tests are described in this section:

The following sections are in this chapter:

and functions are not compromised.

  • “Electrical Safety Tests” on page 89
  • “Resident Self Tests” on page 107

6 Performance Tests

  • “PC Diagnostics” on page 105

its safe and accurate operation.

NOTES

Safety Tests

Electrical

Page 90 of the Philips Patient Monitor (M2540) user manual — Resistance Test
Page 90 · click to enlarge
Page 90
  • The ECG Lead Isolation Leakage test checks ECG leads by applying voltage from the mains to

applying mains voltage to the leads, and checking the transducer for leakage. See “Transducer

step in any repair or upgrade, to maintain agency approved status. The safety analyzer is

also useful in detecting abnormalities of line voltage, grounding, and total current loads.

Performance TestsElectrical Safety Tests

the leads, to assure that the system is isolated from such voltages. See “ECG Lead Isolation

  • The Transducer Isolation Leakage Current test checks a transducer for electrical leakage by

measurements can be performed with multifunction instruments like the HP 3469A multimeter.

The limits referenced in these tests are prescribed by UL, CSA, and IEC. Local regulations may

  • The Transducer Leakage Current test checks a transducer for electrical leakage while it is

transducer, which is tested for leakage. See “Transducer Leakage Current Test Procedure”

  • The ECG Lead Leakage Current test checks ECG leads connected to the optional Physio

module for electrical leakage. See “ECG Lead Leakage Current Test (Source)” on page 94.

connected to the ultrasound system. The system sends normal operating voltages to the

All tests can be performed using commercially available safety analyzer test equipment. Basic

Figure 6-1 on page 91 shows the basic electrical concept of the ground resistance test. Use a

This test checks the entire system for electrical resistance between the chassis and ground.

safety analyzer and complete the procedure that follows the figure.

Follow the instructions of the analyzer manufacturer.

Isolation Leakage Current Test” on page 102

Leakage Current Test” on page 96.

require additional tests.

on page 100.

Resistance Test

Chassis to

Ground

Page 91 of the Philips Patient Monitor (M2540) user manual — Chassis to Ground Resistance Test Procedure
Page 91 · click to enlarge
Page 91
3. Plug the analyzer into an available AC wall outlet. Plug the ultrasound system power plug into the test receptacle on the analyzer.
Performance TestsElectrical Safety Tests

10 amps. Test currents in excess of 1 amp will likely exceed the impedance limit specification, and can damage the RFI protection finger contacts. Exposed metal on the transducer assembly,

2. Set the mode on the analyzer to measure resistance in the power cord.
Philips Patient Monitor (M2540) — Chassis to Ground Resistance Test Procedure diagram (page 91)
Fig. 33 · Chassis to Ground Resistance Test Procedure

including the connector, are for RFI and are not safety grounds.

To test the system for resistance between chassis and ground:

➤Chassis to Ground Resistance Test Procedure

Figure 6-1 Chassis to Ground Resistance Test Diagram

1. Inspect the power cord for cracks and wear.

R = 200 milliohms maximum

LIMITUL, CSA and IEC:

300e036 · (resistance)

Page 92 of the Philips Patient Monitor (M2540) user manual — Ground Wire
Page 92 · click to enlarge
Page 92

shows the basic electrical concept of the test. Use a safety analyzer and complete the procedure

Performance TestsElectrical Safety Tests
5. Read chassis ground resistance in milliohms. Flex the ultrasound power cord during the test to detect intermittent changes in resistance value.

voltage drop across the entire length of the power cord and to the system chassis. The reading

6. Record the highest resistance value measured in step 5 . Check that the highest resistance value is within the limit specified in Figure 6-1 on page 91 .

A comprehensive ground impedance measurement can be performed using the safety analyzer.

This test checks the entire system for leakage on chassis ground wires. Figure 6-2 on page 93

The impedance test drives a load current through the ground wire while measuring the AC

4. Make the appropriate connections between the analyzer and the ground lug on the ultrasound system’s I/O panel.

ground connection is open, do not touch the M2540A chassis or the patient cable.

primary wiring.

that follows the figure.

will be in milliohms.

of the system.

Ground Wire

Current Test

Leakage

Page 93 of the Philips Patient Monitor (M2540) user manual — LIMITS:
Page 93 · click to enlarge
Page 93
2. Plug the analyzer into an available wall outlet. Plug the ultrasound system power plug into the test receptacle on the analyzer. Turn on the ultrasound system’s power by starting the
Performance TestsElectrical Safety Tests
3. Take ground wire leakage current measurements in an open ground condition, with both normal and reverse polarity. Record the highest value, and compare it to the limit specified

simultaneously for Single Fault condition

Philips Patient Monitor (M2540) — Ground Wire Leakage Current Test Procedure diagram (page 93)
Fig. 34 · Ground Wire Leakage Current Test Procedure

1000 microamperes Single Fault condition 7ASW025

➤Ground Wire Leakage Current Test Procedure

Figure 6-2 Ground Wire Leakage Current Test Diagram

1. Set the mode on the analyzer to detect leakage.

1000 microamperes Single Fault condition

for Normal condition in Figure 6-2 on page 93.

To test the system’s ground wiring for leakage:

ULI = 300 microamperes Normal condition,
CSA and IECI = 500 microamperes Normal condition,

internal PC.

LIMITS:

(current)

Page 94 of the Philips Patient Monitor (M2540) user manual — Current Test
Page 94 · click to enlarge
Page 94

Figure 6-3 on page 95 shows the basic electrical concept of the ECG Lead Leakage Current test.

Performance TestsElectrical Safety Tests
4. Take ground wire leakage current measurements in an open ground and open neutral condition, with both normal and reverse polarity. Record the highest value, and compare it

A reading higher than acceptable limits can indicate a problem with the power cord, with its

This test checks ECG leads connected to the optional Physio module for electrical leakage.

ground connection is open, do not touch the M2540A chassis or the patient cable.

to the limit specified for Single Fault condition in Figure 6-2 on page 93.

Use a safety analyzer and complete the procedure that follows the figure.

associated connections, or with the power transformer.

Current Test

ECG Lead

Leakage

(Source)

Page 95 of the Philips Patient Monitor (M2540) user manual — 7ASW026
Page 95 · click to enlarge
Page 95
2. Plug the analyzer into an available AC wall outlet. Plug the ultrasound system power plug into the test receptacle on the analyzer.
4. Connect all ECG leads to the ultrasound system and to the appropriate jacks on the analyzer.
Performance TestsElectrical Safety Tests
Philips Patient Monitor (M2540) — ECG Lead Leakage Current Test Procedure diagram (page 95)
Fig. 35 · ECG Lead Leakage Current Test Procedure

7ASW026

3. Power the ultrasound system by starting its internal PC.

Neutral wire open (ground closed) for second Single Fault condition

➤ECG Lead Leakage Current Test Procedure

Figure 6-3 ECG Lead Leakage Current Test Diagram

1. Set the analyzer to test ECG leads.

50 microamperes Single Fault condition

I = 10 microamperes Normal condition

For

Page 96 of the Philips Patient Monitor (M2540) user manual — Current Test
Page 96 · click to enlarge
Page 96
5. Take ECG lead leakage current measurements with both normal and reverse polarity. Check that the highest value is within the limit specified for Normal condition in Figure 6-3 on
6. Take ECG lead leakage current measurements in open ground condition, with both normal and reverse polarity. Check that the highest value is within the limit specified for Single Fault

condition in Figure 6-3. Write down the highest value; this is the value for the first single fault

7. Take ECG lead leakage current measurements in open neutral condition, with both normal and reverse polarity. Check that the highest value is within the limit specified for Single Fault
8. Record the highest current value measured in step 6 and step 7 . Check that the highest value is within the limit specified for Single Fault condition in Figure 6-3 .

concept of the ECG Lead Isolation Leakage Current test. Use a safety analyzer and complete the

Performance TestsElectrical Safety Tests

condition in Figure 6-3. Write down the highest value; this is the value for the second single

This test checks ECG leads by applying voltage from the mains to the ECG leads, to check that

Failure to meet the specified limits may indicate a fault with the isolation of the Physio module.

the system is isolated from such voltages. Figure 6-4 on page 97 shows the basic electrical

procedure that follows the figure.

fault condition.

condition.

page 95.

Current Test

ECG Lead

Isolation

Leakage

Page 97 of the Philips Patient Monitor (M2540) user manual — 1011E29
Page 97 · click to enlarge
Page 97
2. Plug the analyzer into an available AC wall outlet. Plug the ultrasound system power plug into the test receptacle on the analyzer.
Performance TestsElectrical Safety Tests

line voltage. Do not touch the chassis or the ECG cable while performing the test. Keep the

1011E29

ECG cable at least 20 cm from any grounded or conductive surfaces.

Philips Patient Monitor (M2540) — 1011E29 diagram (page 97)
Fig. 36

➤ECG Lead Isolation Leakage Current Test Procedure

3. Power up the ultrasound system by starting its internal PC.

Figure 6-4 ECG Lead Isolation Leakage Current Test Diagram

1. Set the analyzer’s mode to test ECG isolation.
UL, IEC, and CSAI = 50 microamperes RMS

open neutral conditions.

Power cord connected to AC power.

ECG patient

Page 98 of the Philips Patient Monitor (M2540) user manual — Transducer
Page 98 · click to enlarge
Page 98
6. Connect all ECG leads to the appropriate jacks on the analyzer and perform a lead isolation test. Record the reading in microamperes.

patient safety, it is very important to verify the integrity of the insulating layers of all transducers.

When performing the safety tests, saline solution in a container is used as a conductive medium

about the transducer tests. This is supplemental information, and is not required to complete the

Performance TestsElectrical Safety Tests
7. Subtract the step 5 correction factor from the step 6 reading to get an accurate isolation leakage measurement. Record this value. Check that the value is within the limit specified in
“Transducer Safety TestingTest Setup and Theory” on page 52 includes supporting information
5. In normal polarity, perform a lead isolation test. Record the reading in microamperes. This reading is the correction factor used in step 7 .
4. Connect ECG leads to the ultrasound system, but do not connect the ECG leads to the analyzer yet.

Tests Two electrical leakage current tests for transducers are described in this section. To insure

(see Figure 4-4 on page 53). The solution penetrates any cracks or holes in the transducer

insulation and provides an electrical path between the submerged lead wire and the inner

transducer is a TEE, check that its steering controls are working properly.

reading. Refer to the analyzer’s documentation for more information.

Figure 6-4 on page 97.

transducer shield.

tests.

Transducer

Page 99 of the Philips Patient Monitor (M2540) user manual — Transducer Leakage Current Test (Source)
Page 99 · click to enlarge
Page 99
Performance TestsElectrical Safety Tests

system. The system sends normal operating voltages to the transducer, and leakage is measured

  • Saline solution. If saline is not available, mix 27 grams of table salt in 3 liters of tap water.

This test checks a transducer for electrical leakage while it is connected to the ultrasound

using a safety analyzer. Figure 6-5 on page 100 shows the basic electrical concept of the

Transducer Leakage Current test. Complete the procedure that follows the figure.

The following tools are required for transducer safety tests:

Transducer Leakage Current Test (Source)

  • Safety analyzer
  • ECG lead wire
  • Saline holder
Page 100 of the Philips Patient Monitor (M2540) user manual — 7ASW027
Page 100 · click to enlarge
Page 100
2. Plug the analyzer into an available AC wall outlet. Plug the ultrasound system power plug into the test receptacle on the analyzer.
Performance TestsElectrical Safety Tests
4. Plug the transducer to be tested into the ultrasound system. Connect an ECG lead wire to the appropriate jack on the analyzer.

ECG

Neutral wire open (ground closed) for second Single Fault condition Ground wire open for first Single Fault condition, lead

500 microamperes Single Fault condition I = 100 microamperes Normal condition,

7ASW027

Philips Patient Monitor (M2540) — 7ASW027 diagram (page 100)
Fig. 37
3. Power up the ultrasound system by starting its internal PC.

For type transducers:

➤Transducer Leakage Current Test Procedure

Figure 6-5 Transducer Leakage Current Test Diagram

1. Set the analyzer mode to test ECG leads.

50 microamperes Single Fault condition I = 10 microamperes Normal condition, For type transducers:

LIMITSUL, IEC, and CSA:

wire · Hot -

Page 101 of the Philips Patient Monitor (M2540) user manual — Transducer Leakage Current Test Procedure
Page 101 · click to enlarge
Page 101
  • For TEE transducers, submerge all of the flexible shaft that would normally enter the

saline to contact its control handle. Liquid inside the control handle causes corrosion and failure.

page 100. (Match the transducer type symbol on the transducer connector or cable with the

Performance TestsElectrical Safety Tests
9. Read the current with an open neutral condition imposed, in both normal and reverse polarity. Write down the highest value; this is the value for the second single fault condition.
5. Fill a test container with enough saline to completely cover the appropriate parts of the transducer. Insert the exposed end of the ECG lead into the saline, and carefully insert the
7. Read the current in both normal and reverse polarity, and record the highest value. Check that the highest value is within the limit specified for Normal condition in Figure 6-5 on
  • For transthoracic and endocavity transducers, submerge the head and 5 cm of the
8. Read the current with an open ground condition imposed, in both normal and reverse polarity. Write down the highest value; this is the value for the first single fault condition.
10. Check that the highest current value measured in step 8 and step 9 is within the limit specified for Single Fault condition in Figure 6-5 .

body of the patient: 100 cm for adult TEE, 60 cm for pediatric TEE.

6. Set the analyzer to read leakage current in microamperes.

cable. Be careful not to submerge the connector.

symbol in the figure.)

transducer:

Page 102 of the Philips Patient Monitor (M2540) user manual — Transducer Isolation Leakage Current Test
Page 102 · click to enlarge
Page 102

Values exceeding the limits may indicate a fault in the transducer housing or the cable sheath.

To locate a fault precisely, repeat the measurement while slowly inserting the transducer into

of the cable. Their leakage currents can measure within the specified limits even if there

Performance TestsElectrical Safety Tests

Current test. The transducer is submerged in a container of saline solution. Mains AC voltage is

11. In normal polarity with an open ground condition imposed, measure ground wire leakage (see “Ground Wire Leakage Current Test” on page 92 ). Compare this to the highest value

Figure 6-6 on page 103 shows the basic electrical concept of the Transducer Isolation Leakage

If the transducer leakage is greater than 80% of the chassis leakage, there is a fault in the

introduced into the saline solution, and leakage current in the transducer is measured using a

transducer. To locate the fault, slowly insert the transducer in the saline, as in step 10.

the saline. The aberrant reading appears when the fault in the cable enters the saline.

for transducer leakage current, Single Fault condition obtained in step 10.

is a fault in the outer skin. Step 11 tests for this condition.

safety analyzer. Complete the procedure that follows the figure.

Transducer Isolation Leakage Current Test

Page 103 of the Philips Patient Monitor (M2540) user manual — ECG
Page 103 · click to enlarge
Page 103
Performance TestsElectrical Safety Tests

contact with the line voltage. Do not touch the chassis, or the ECG or transducer cables while

performing the test. Keep the ECG and transducer cables at least 20 cm from any grounded or

ECG

Philips Patient Monitor (M2540) — ECG diagram (page 103)
Fig. 38

➤Transducer Isolation Leakage Current Test Procedure

Figure 6-6 Transducer Isolation Leakage Current Test Diagram

open neutral conditions.

1. Set the analyzer to test ECG.

conductive surfaces.

wire · 1101e30

Page 104 of the Philips Patient Monitor (M2540) user manual — Transducer Isolation Leakage Current Test Procedure
Page 104 · click to enlarge
Page 104
2. Plug the analyzer into an available AC wall outlet. Plug the ultrasound system power plug into the test receptacle on the analyzer.
  • For TEE transducers, submerge all of the flexible shaft that would normally enter the

saline to contact its control handle. Liquid inside the control handle causes corrosion and failure.

Performance TestsElectrical Safety Tests
7. Measure the isolation leakage current of the transducer. Record the value, and check that it is within the limit specified in Figure 6-6 on page 103 .
  • For transthoracic and endocavity transducers, submerge the head and 5 cm of the
5. Fill a test container with enough saline to completely cover the appropriate parts of the transducer.
6. Insert the exposed end of the ECG lead wire into the saline, then carefully insert the transducer:

body of the patient: 100 cm for adult TEE, 60 cm for pediatric TEE.

3. Power up the ultrasound system by starting its internal PC.
4. Plug the transducer to be tested into the ultrasound system.

cable. Be careful not to submerge the connector.

Page 105 of the Philips Patient Monitor (M2540) user manual — IBM Drive
Page 105 · click to enlarge
Page 105
Performance TestsPC Diagnostics
1. Change the VGA display height adjustment to show the whole screen. See “Adjusting VGA Vertical Size” on page 128 .
2. Change the boot sequence so the floppy disk is the first item. (See “Changing the Boot Sequence” on page 130 .)

PC problems can often be diagnosed using the IBM Drive Fitness test or Checkit Pro.

You must make a bootable floppy disk to use the Drive Fitness test.

1. Download the bootable version of the DFT setup program from http://www.storage.ibm.com/hdd/support/download.htm
6. Insert a blank formatted 1.44 MB diskette when prompted.

➤Testing PC Functions Using the Drive Fitness Test

3. Click the Browse button to locate the setup program.

http://www.storage.ibm.com/hdd/technolo/dft/dftnew.htm

3. Insert the DFT disk in the floppy disk drive.

➤Creating a DFT Bootable Floppy Disk

Fitness Test The Drive Fitness test (DFT) is described at:

into a temporary directory on your PC.

4. Click the filename, then click Open .
5. Click OK to run the setup program.
2. Click Start , then select Run .
5. Select ATA support .
4. Reboot the system

Diagnostics

IBM Drive

PC

Page 106 of the Philips Patient Monitor (M2540) user manual — Checkit Pro
Page 106 · click to enlarge
Page 106
Performance TestsPC Diagnostics
1. Change the VGA display height adjustment to show the whole screen. See “Adjusting VGA Vertical Size” on page 128 .
6. When the CheckIt Pro testing is complete, enter the BIOS again, and make the CD-ROM drive the first item in the boot sequence. (See “Changing the Boot Sequence” on page 130 .)
2. Change the boot sequence so the floppy disk is the first item. (See “Changing the Boot Sequence” on page 130 .)

Checkit Pro Test PC functions using the CheckIt Pro software on the Diagnostic floppy disk.

http://service.boulder.ibm.com/storage/hddtech/DFT32-V300USER-DOC.pdf

first. (See “Changing the Boot Sequence” on page 130.)

3. Insert the Diagnostic disk in the floppy disk drive.

➤Testing PC Functions Using Checkit Pro

For detailed information on using DFT, go to:

6. Follow the onscreen instructions.
5. Run the CheckIt Pro application.
4. Restart the system
Page 107 of the Philips Patient Monitor (M2540) user manual — Resident Self
Page 107 · click to enlarge
Page 107
Performance TestsResident Self Tests

Customers or service personnel who log on with a valid user name and password can create and

The M2540A Resident Self Tests (RST) are a suite of tests that verify system hardware, using no

The stand-alone RST application provides access to all the tests. RST also provides information

The logon window also includes a check box that skips the logon dialog the next time the RST

The RST suite consists of tests for custom M2540A hardware and software. The Resident Self

Tests also exercise individual subsystems such as the acoustic processor and the beam former

information, test pass and fail results, and errors can be viewed, saved, and printed by the RST

application opens with a logon window, which asks you to enter a user name and a password.

Availability The RST application is available whenever the ultrasound application is running. You can also

application is launched. Only service personnel can enable that option, and it only applies for

Logging On To launch the RST application, see “Launching the RST Application” on page 113. The RST

launch RST as the system powers up. When it starts, the RST application stops the main

24 hours. After 24 hours, the logon dialog shows at launch, even if the option is selected.

on hardware and software, and service information on the ultrasound system. System

modify service log entries. See “Using the Service Log” on page 125.

If RST is installed on the system, logging on is not required.

external instrumentation.

ultrasound application.

application.

processor.

Resident Self

Tests

Page 108 of the Philips Patient Monitor (M2540) user manual — User Interface
Page 108 · click to enlarge
Page 108
Performance TestsResident Self Tests

User Interface You can control the RST software using the system control panel’s keyboard and trackball. When

hierarchy of supported functions. A list view is on the right side. The list view displays information

using the system control panel, the two keys just above the trackball perform the functions of

mouse buttons. The Enter key acts as a left mouse button, and the Select key acts as a right

A tab view is on the left side of the RST window. On each tab is tree control that displays a

Philips Patient Monitor (M2540) — User Interface diagram (page 108)
Fig. 39 · User Interface

related to items selected in a tree control.

Figure 6-7 shows the RST window.

Figure 6-7 RST Window

mouse button.

Page 109 of the Philips Patient Monitor (M2540) user manual — Information
Page 109 · click to enlarge
Page 109
Performance TestsResident Self Tests
  • The Info tab shows software, hardware, and system information. See “Information Types” on
  • The Log tab displays errors and test parameter data in the event viewer and displays service

log on the Log tab, for service personnel to record information about the system. See “Using

Choosing a tree item and pressing the Select key opens a drop-down list that contains Save Info

  • The Test tab invokes various test modes and displays test status and results. See “Tests” on

log entries in the service log. See “Viewing Error Logs” on page 119. There is also a service

and Print Info functions. Information can be saved to a text file or printed on a connected plain

institution name, installed software and hardware options, and installed peripherals. Figure 6-7,

The software information provided includes the revision number and the revision date and time

of the operating system, the main ultrasound application, and the common off-the-shelf (COTS)

Types The RST application provides information on the hardware and the software of the ultrasound

system. You can select the Info tab and choose specific information types in the tree control.

The RST application provides system information that includes the system serial number, the

There are three tabs in the RST window:

the Service Log” on page 125.

on page 108 illustrates this display.

System Information

paper printer.

page 109.

software.

Information

page 111

Page 110 of the Philips Patient Monitor (M2540) user manual — Hardware Information
Page 110 · click to enlarge
Page 110
Performance TestsResident Self Tests

The hardware information provided includes information about the custom M2540A boards and

Philips Patient Monitor (M2540) — Hardware Information diagram (page 110)
Fig. 40 · Hardware Information

transducer connector modules, identified as shown in the following figure:

connector modules installed.

Figure 6-8 Hardware Information Display

Hardware Information

Page 111 of the Philips Patient Monitor (M2540) user manual — Available Tests
Page 111 · click to enlarge
Page 111
Performance TestsResident Self Tests

five minutes. The test requires no operator interaction while executing. See “Running a Basic

  • Basic Test : A basic test of custom M2540A hardware. The basic test executes in less than

The following tests are available when you log on using a valid service name and password:

Philips Patient Monitor (M2540) — Tests diagram (page 111)
Fig. 41 · Tests

Test” on page 114 for instructions on running basic tests.

Tests The following figure shows the RST Test display.

Available Tests

Figure 6-9 RST Test Display

Page 112 of the Philips Patient Monitor (M2540) user manual — Testing Options
Page 112 · click to enlarge
Page 112
Performance TestsResident Self Tests

selected tests run until completed. Any errors occurring during the tests are recorded in the

operator interaction while it executes. The test verifies detailed system operation with fault

The left-hand panel under the RST Test tab contains three testing options and the Run button

  • Extended Test : This is a comprehensive test of M2540A hardware. The test requires no
  • Interactive Test: These tests verify operation of the system’s controls, and require your
  • Repeat Test Count defines the number of time tests are repeated. The default is one. A

participation. Included are tests of the display and system control panel. See “Running an

isolation to the board level. This test reports only failures that adversely impact system

  • Stop On Errors stops all tests when an error occurs. If this option is not enabled, the

performance. See “Running an Extended Test” on page 115 for instructions on running

that begins a selected test. When an interactive test begins, the Run button is grayed out.

Interactive Test” on page 118 for instructions on running interactive tests.

Disconnect all transducers before beginning any RST testing.

  • Save Parameter Data saves the measured test parameters.

value of zero runs tests continuously.

The three test options are:

Testing Options

extended tests.

error log.

Page 113 of the Philips Patient Monitor (M2540) user manual — Service Log
Page 113 · click to enlarge
Page 113
Performance TestsResident Self Tests

Test Results The test results are shown in the list view on the right-hand side of the RST window. The display

shows the name of each test, the components it tested, and the number of times the component

You can enter notes in the service log section of the event viewer, to record service procedures

  • Press the key combination Ctrl+Alt+R while the ultrasound splash screen is on the display.

To view details of the test results, click the Log tab, and select the information type from the

Event Viewer tree. For information on errors generated during tests, see “RST Test Log” on

You can launch the RST application as the PC boots up or while the ultrasound application is

you perform or problems you encounter. See “Using the Service Log” on page 125.

When it launches, the RST application closes the ultrasound application.

When the tests are finished, the following message box opens:

To launch the RST application when the system is fully booted:

To launch the RST application during system boot-up:

running. The procedure is the same in both cases.

2. Press the key combination Ctrl+Alt+R .

➤Launching the RST Application

1. Launch the ultrasound application.

passed and failed the test.

Philips Patient Monitor (M2540) — Test Results diagram (page 113)
Fig. 42 · Test Results

Service Log

page 121.

Page 114 of the Philips Patient Monitor (M2540) user manual — Closing the RST Application
Page 114 · click to enlarge
Page 114
Performance TestsResident Self Tests
5. To run the test a specific number of times, enter the number in the Repeat Test Count: field. To run the test continuously, enter zero in the field.

The basic test checks custom M2540A hardware. The basic test executes in less than five

1. Launch the RST application. (See “Launching the RST Application” on page 113 .)

You can close the RST application using either of the following two methods:

  • Click the Close button at the upper right corner of the RST window.

When the RST application closes, a window asks if you want to reboot.

minutes and requires no operator interaction while executing.

7. When the testing is done, click OK in the message box.
2. Log on using a valid service name and password.
3. Click the Test tab in the main RST window.
  • Press the Alt and Report keys together.
4. Click Basic Test in the tree view.

➤Closing the RST Application

6. Click Run . Testing begins.

➤Running a Basic Test

Philips Patient Monitor (M2540) — Closing the RST Application diagram (page 114)
Fig. 43 · Closing the RST Application

To run a basic test:

Page 115 of the Philips Patient Monitor (M2540) user manual — Running an Extended Test
Page 115 · click to enlarge
Page 115
Performance TestsResident Self Tests

The extended tests check the system’s hardware to the board level, and do not require your

1. Launch the RST application. (See “Launching the RST Application” on page 113 .)
4. Click Extended Test in the tree view, to show the available extended tests.
2. Log on using a valid service name and password .
3. Click the Test tab in the main RST window.

participation after they are started.

➤Running an Extended Test

To run an extended test:

Page 116 of the Philips Patient Monitor (M2540) user manual — Running an Extended Test
Page 116 · click to enlarge
Page 116
Performance TestsResident Self Tests
5. If you want to run specific tests, click the names of those tests. (See “Test Modules” on page 6-117 .) To select all the extended tests, leave Extended Test selected.
6. To run the test a specific number of times, enter the number in the Repeat Test Count: field. To run the test continuously, enter zero in the field.
Philips Patient Monitor (M2540) — Running an Extended Test diagram (page 116)
Fig. 44 · Running an Extended Test
8. When the testing is done, click OK in the message box.
7. Click Run . Testing begins.

Figure 6-10 RST Extended Tests

Page 117 of the Philips Patient Monitor (M2540) user manual — Test Modules
Page 117 · click to enlarge
Page 117
Performance TestsResident Self Tests

using the tree control in the left side of the RST window. As an example, Figure 6-11 shows the

Each subsystem test consists of test modules, and each test module has submodules. You can

select individual modules or submodules to run separate from the other tests in its group by

BP/AP subsystem test’s Demodulator test module and the module’s submodules.

Philips Patient Monitor (M2540) — Test Modules diagram (page 117)
Fig. 45 · Test Modules

Figure 6-11 Demodulator Tests (Example)

Test Modules

Page 118 of the Philips Patient Monitor (M2540) user manual — Running an Interactive Test
Page 118 · click to enlarge
Page 118
Performance TestsResident Self Tests

display, in its Category column, and in the Status window at the lower left of the window. See

  • To end a keyboard test, right-click the test name, and select Stop from the pop-up

The two-digit codes that precede the submodule names are used in the Application Test Log

The interactive tests check the operation of the system’s controls. They require you to make

1. Launch the RST application. (See “Launching the RST Application” on page 113 .)
4. Click Interactive Test in the tree view, to show the available interactive tests.
7. Follow the instructions that the RST application displays.
2. Log on using a valid service name and password.
  • To end a display test, press the THI key.

selections and operate controls as they execute.

3. Click the Test tab in the main RST window.
5. Select the test you want to run.

➤Running an Interactive Test

8. To end a test, do the following:

“RST Test Log” on page 121.

6. Click Run . Testing begins.

To run an interactive test:

Philips Patient Monitor (M2540) — Running an Interactive Test diagram (page 118)
Fig. 46 · Running an Interactive Test

menu.

Page 119 of the Philips Patient Monitor (M2540) user manual — Ultrasound Application Log
Page 119 · click to enlarge
Page 119
Performance TestsResident Self Tests

The logs list errors in chronological order, with the most recent errors at the top of the list. You

The RST application logs errors during operation of the system and when you run a test. View

the logs by clicking Event Viewer under the Log tab of the RST window and selecting the log

This log includes run-time errors encountered by both the ultrasound application and the RST

application. The log lists the date and time the error occurred, the name of the error, the

can view more detail of an error by double-clicking the entry in the list view.

Philips Patient Monitor (M2540) — Viewing Error Logs diagram (page 119)
Fig. 47 · Viewing Error Logs

The Event Viewer includes application logs, a security log, and a system log.

you want to see. Figure 6-12 shows the RST log display.

Ultrasound Application Log

➤Viewing Error Logs

Figure 6-12 RST Log Display

Page 120 of the Philips Patient Monitor (M2540) user manual — Clearing the Ultrasound Application Log
Page 120 · click to enlarge
Page 120
Performance TestsResident Self Tests

applications and subsystems affected, and other information. Figure 6-13 shows the ultrasound

Philips Patient Monitor (M2540) — Clearing the Ultrasound Application Log diagram (page 120)
Fig. 48 · Clearing the Ultrasound Application Log

To clear the Ultrasound Application Log, complete the following procedure:

➤Clearing the Ultrasound Application Log

3. Click Yes in the confirmation dialog box.
1. Right-click on Ultrasound App Log .
2. In the pop-up menu, select Clear Log .

Figure 6-13 Ultrasound Application Log Display

application log display.

Page 121 of the Philips Patient Monitor (M2540) user manual — RST Test Log
Page 121 · click to enlarge
Page 121
Performance TestsResident Self Tests

This log includes parameters of past RST tests and errors encountered while running the tests.

Philips Patient Monitor (M2540) — RST Test Log diagram (page 121)
Fig. 49 · RST Test Log

Figure 6-14 shows the Event Viewer application RST test log.

Figure 6-14 RST Application Test Log Display

RST Test Log

Page 122 of the Philips Patient Monitor (M2540) user manual — RST Log Codes
Page 122 · click to enlarge
Page 122
Performance TestsResident Self Tests

To find the meaning of the numeric codes, use the tree control in the RST window’s left side as

The example indicates that the A/D converter (submodule 02 of module 01) of the display

The Application Test Log’s Category column contains coded information. The code is in the

  • nn is the module number of the tested module.
  • DP is the display processor subsystem.
  • BP is the beam processor subsystem.
1. Right-click RST Test Log in the tree view.
3. Click Yes in the confirmation dialog box.
2. In the pop-up menu, select Clear Log .

described in “Test Modules” on page 117.

  • NN is the submodule number.

➤Clearing the RST Test Log

  • SS is the subsystem tested.

processor was tested.

RST Log Codes

ExampleDP0102

following format:

Where:

SSnnNN

Page 123 of the Philips Patient Monitor (M2540) user manual — Test Result Details
Page 123 · click to enlarge
Page 123
Performance TestsResident Self Tests
1. In the right-hand side of the Event Viewer window, double-click the event you want detailed information on.
2. Click the appropriate Data: radio button to change the data display type.

To view detailed test results, open the Event Properties window as follows:

Philips Patient Monitor (M2540) — Test Result Details diagram (page 123)
Fig. 50 · Test Result Details

Figure 6-15 Event Properties Window (Example)

The Event Properties window opens.

➤Viewing Test Details

Test Result Details

Page 124 of the Philips Patient Monitor (M2540) user manual — System Log
Page 124 · click to enlarge
Page 124
Performance TestsResident Self Tests

The Event Viewer system log includes hardware errors filtered from the application run-time

error log and the RST test log. Figure 6-16 shows the Event Viewer system log.

Philips Patient Monitor (M2540) — System Log diagram (page 124)
Fig. 51 · System Log
1. Right-click System Log in the tree view.
3. Click Yes in the confirmation dialog box.
2. In the pop-up menu, select Clear Log .

➤Clearing the System Log

Figure 6-16 System Log Display

System Log

Page 125 of the Philips Patient Monitor (M2540) user manual — Using the
Page 125 · click to enlarge
Page 125
Performance TestsResident Self Tests

You can enter information in the service log that is displayed under the Log tab. You must log on

to the RST application using a valid service name and password to make an entry in the service

3. Right-click Service Log on the tree (use the Select key on the system control panel). A pop-up menu appears.
5. Enter your name in the top two fields, and the service information in the Comments: field.
4. Click Add Service Entry on the pop-up menu. The Add Service Entry window displays.

Service Log The service log on the Log tab’s tree is a list of comments entered by service personnel.

1. Launch the RST application. (See “Launching the RST Application” on page 113 .)
2. Log on using a valid service name and password.
6. Click OK . Your entry appears in the service log.

➤Making an Entry in the Service Log

To make a new entry in the service log:

Philips Patient Monitor (M2540) — Using the Service Log diagram (page 125)
Fig. 52 · Using the Service Log
Philips Patient Monitor (M2540) — Using the Service Log diagram (page 125)
Fig. 53 · Using the Service Log

log.

Using the

Page 126 of the Philips Patient Monitor (M2540) user manual — Modifying an Entry in the Service Log
Page 126 · click to enlarge
Page 126
Performance TestsResident Self Tests
4. Double-click the log entry you want to modify. The Change Service Entry window displays.

You can change the information in an existing service log entry. You must log on to the RST

application using a valid service name and password to modify an entry in the service log.

1. Launch the RST application. (See “Launching the RST Application” on page 113 .)
7. Right-click Service Log and select Save Log to File to save the modification.
Philips Patient Monitor (M2540) — Modifying an Entry in the Service Log diagram (page 126)
Fig. 54 · Modifying an Entry in the Service Log
7. Right-click Service Log and select Save Log to File to save the entry.
2. Log on using a valid service name and password.

➤Modifying an Entry in the Service Log

5. Click in the entry field and modify the text.
3. Click Service Log on the tree.

To modify a service log entry:

6. Click OK .
Page 127 of the Philips Patient Monitor (M2540) user manual — Adjustment
Page 127 · click to enlarge
Page 127
AdjustmentsMonitor Adjustment

available. Submenus of the OSD Main Menu control the size and geometry of the display and set

2. Press the two right-hand monitor buttons, and hold them in while reconnecting the power cord.

The four buttons on the front of the monitor adjust brightness and contrast. When the two

right-hand buttons are pressed together as the monitor powers up, the OSD Main menu is

the monitor’s signal input voltage (see “Monitor Signal Voltage” on page 62).

Pressing the two right-hand buttons together degausses the CRT.

1. Disconnect the power cord from the rear of the monitor.
3. Press the two center monitor buttons.

OSD Main Menu ➤Opening the OSD Main Menu

The OSD Main Menu displays.

7 Adjustments

Adjustment

Monitor

Page 128 of the Philips Patient Monitor (M2540) user manual — Adjusting VGA
Page 128 · click to enlarge
Page 128
AdjustmentsMonitor Adjustment
4. Press the two right-hand monitor buttons, and hold them in while reconnecting the power cord.

Vertical Size For viewing BIOS settings mode, you must adjust the VGA vertical size. This adjustment does

  • Contrast up (far-right button) moves up to the previous OSD menu item or increases the
2. As the system boots up, press the Patient key repeatedly, until the BIOS screen opens.
  • Contrast down moves down one OSD Menu item or decreases the value of the selected

The four buttons on the front of the monitor control all the OSD functions, as follows:

  • Brightness up + Brightness down e xits a menu without saving any changes.
  • Brightness down (far-left button) applies the selected menu item.
3. Disconnect the power cord from the rear of the monitor.
5. Press the two center monitor buttons.

➤Adjust the VGA Vertical Size

Navigating the OSD Menus

value of the selected parameter.

not affect the normal display mode.

The OSD Main menu displays.

1. Reboot the system.

parameter.

Adjusting VGA

Page 129 of the Philips Patient Monitor (M2540) user manual — Navigating the OSD Menus
Page 129 · click to enlarge
Page 129
AdjustmentsMonitor Adjustment
12. Use the two right-hand monitor buttons to select SAVE & RETURN in the SIze & Position menu.
6. Use the two right-hand monitor buttons to scroll up or down in the OSD Main menu and select SIZE & POSITION .
10. Use the two right-hand monitor buttons to adjust the vertical size so the bottom of the display is visible.
14. In the OSD Main menu, use the two right-hand monitor buttons to select EXIT OSD .
8. Use the two right-hand monitor buttons to select VERTICAL SIZE .

The vertical size adjustment bar displays.

The vertical size adjustment bar closes.

The Size & Position menu displays.

7. Press the far left monitor button.
9. Press the far left monitor button.
11. Press the far left monitor button.
13. Press the far left monitor button.

The SIze & Position menu closes.

Philips Patient Monitor (M2540) — Navigating the OSD Menus diagram (page 129)
Fig. 55 · Navigating the OSD Menus
Page 130 of the Philips Patient Monitor (M2540) user manual — Changing the
Page 130 · click to enlarge
Page 130
AdjustmentsMonitor Adjustment
1. Change the VGA display height adjustment to show the whole screen. See “Adjusting VGA Vertical Size” on page 128 .
11. Press the power switch to shut down the system. (Don’t let it reboot from the dialog box.)
3. As the system boots up, press the Patient key repeatedly, until the BIOS screen opens.
5. Use the arrow keys to select Boot Device Priority . Press Enter .
12. After 15 seconds, press the power switch to start the system.
  • Select Removable Dev. for the floppy disk drive.
  • Select ATAPI CD-ROM for the CD-ROM drive.
6. Use the arrow keys to select the drive to boot from:
8. Press the THI key to return to the Device menu.
4. Use the arrow keys to select Boot . Press Enter .
9. Use the arrow keys to select Exit . Press Enter .
10. Press Enter to confirm the changes.
15. Press the far left monitor button.

Boot Sequence ➤Change the Boot Sequence

The Device menu opens.

2. Reboot the system.
7. Press Enter .

Changing the

Page 131 of the Philips Patient Monitor (M2540) user manual — 8 Preventive Maintenance
Page 131 · click to enlarge
Page 131

The air filter resides inside the service panel at the right front of the lower enclosure, in front of

the fan. If the filter becomes very dirty, it can impede airflow, which may cause overheating of the

1. Remove the service panel at the right front of the lower enclosure. See “Front Service Panel” on page 141 .

Preventive Maintenance:

During preventive maintenance, look for conditions that may cause problems, and loose knobs

This section contains procedures for cleaning the M2540A and its components.

3. Wash the filter with a spray of water from the side that faced the fan.

cleaners may cause the colors to fade. Use only mild soap.

8 Preventive Maintenance

2. Slide the filter up and out of the panel’s back.

To remove and clean the air filter:

➤Cleaning the Air Filter

power supply or the e-box.

Safeguards” on page 139.

or worn TGC controls.

Page 132 of the Philips Patient Monitor (M2540) user manual — Cleaning the Trackball
Page 132 · click to enlarge
Page 132

Preventive Maintenance:

4. Dry the filter thoroughly before replacing it in the system. Compressed air can speed drying.
1. Unscrew and remove the bezel that surrounds the trackball in the surface of the system control panel. A paper clip inserted in one of the small holes in the bezel can help to start

Inspect the interior of the e-box for dust (See “Replacing E-box Circuit Boards” on page 151),

4. Clean the two encoder shafts and the bearing surface with an alcohol-dampened pad.
5. Reverse step 1 and step 2 to complete the procedure.
3. Clean the trackball with an alcohol-dampened pad.
5. Reverse step 1 and step 2 to reinstall the trackball.
2. Press up from under the trackball and remove it.

➤Cleaning the Trackball

and vacuum as necessary.

unscrewing the bezel.

➤Cleaning the e-box

To clean the trackball:

material.

Page 133 of the Philips Patient Monitor (M2540) user manual — Degaussing the Monitor
Page 133 · click to enlarge
Page 133

Preventive Maintenance:

protected surface. Use adequate ESD safeguards when performing any service or maintenance

removing them from the system. After removal, make sure the boards lie flat on an ESD

To degauss the monitor, press the two right-hand monitor buttons at the same time.

Perform all RST and system checks to verify performance.

vacuuming. See “ESD Safeguards” on page 139.

procedure. See “ESD Safeguards” on page 139.

➤Degaussing the Monitor

Page 134 of the Philips Patient Monitor (M2540) user manual — PC Won’t
Page 134 · click to enlarge
Page 134
TroubleshootingPC Won’t Boot
3. Unplug the PC power cord from the top outlet on the power supply, and connect the PC to a power source using one of the following methods:
  • If the PC does not start, try replacing its power cord. If the PC still does not boot, it

This section contains tips and procedures for finding the causes of malfunctions of the M2540A.

  • If the PC starts, the power supply is bad. Replace it. See “Power Supply” on page 154 .
2. Remove the service panel from the front of the lower enclosure by pressing down its two latches. A green LED on the left side of the power supply, above the fan, indicates that the

power supply is energized by line voltage from the mains. The LED takes a few seconds to

power supply by jumping the PWR DOWN pin on the BPAP board to chassis ground.

  • Set the PC’s power selector switch to the local voltage, and use the system power

If the M2540A does not start when the PC’s power button is pressed, proceed as follows:

Always check that power cords are plugged in, and inspect them for wear.

  • Plug the PC’s power cord into a known 120 VAC source.

For problems with optional peripherals, see the Peripherals Manual.

1. Confirm that the system power cord is plugged in.

is faulty. See “PC Service” on page 142.

cord to connect the PC to mains.

light when the system is first plugged in.

4. Press the PC’s power button.

9 Troubleshooting

PC Won’t

Boot

Page 135 of the Philips Patient Monitor (M2540) user manual — PC Boots, But
Page 135 · click to enlarge
Page 135
TroubleshootingPC Boots, but Monitor is Blank
4. Move the malfunctioning device’s power plug to the PC power outlet on the power supply.
1. Connect a jumper between chassis ground and the PWR DOWN pin on the BPAP board.
  • If the monitor does not power up, it is faulty. Replace it. See “Replacing the Monitor” on
  • If the device does not power up, try replacing its power cord. If the device still does not
1. Run the RST and other system checks, and check the device’s power cord connections.
  • If the monitor powers up, the power supply is bad. Replace it. See “Power Supply” on

If the M2540A’s internal PC starts, and the monitor functions normally, but one or more

  • If the device powers up, the power supply is bad. Replace it. See “Power Supply” on
5. Reset the PC power selector switch to 120 V, if the switch was moved in step 3 .

If the M2540A’s internal PC starts, but the monitor stays black, proceed as follows:

4. Move the monitor’s power plug to the PC power outlet on the power supply.
  • If the monitor remains black, continue with the following steps.
  • If the monitor now powers up, check the token ring cable.
3. Unplug the PC from the top outlet on the power supply.

peripheral devices don’t work, proceed as follows:

power up, it is faulty. Replace the device.

3. Unplug the PC from the power supply.
2. Shut down the PC.
2. Shut down the PC.

page 154.

page 157.

page 154.

PC Boots, But

PC Boots, but

Peripherals

Don’t Work

Monitor is

Blank

Page 136 of the Philips Patient Monitor (M2540) user manual — Printer Instal-
Page 136 · click to enlarge
Page 136
TroubleshootingPrinter Problems

lation Problems If a connected printer does not show up on the Setup/Peripherals screen, perform the following

2. Make sure that the printer is connected to a live power source, that its USB cable is properly connected at both ends, that the printer is turned on, and that the printer shows no error

HP940 Startup The first time the HP940 printer is powered up, it may cause the system to hunt for more

3. Reboot the M2540A, and see if the printer appears on the Setup/Peripherals screen.
1. Use the scroll bar in the options field to make sure you see all the printers in the list.

If the procedure does not solve the problem, contact your service representative.

peripherals to install. Rebooting the system stops this activity.

6. On the Setup window, click the Peripherals tab.
4. Disconnect the USB cable from the printer.
12. Connect the USB cable to the printer.
7. Click Install Software Drivers .
11. Shut down the M2540A.
5. Press the Setup key.
13. Start the M2540A.
8. Click Printer .
10. Click Close .

indicators.

9. Click OK .

procedure:

Printer Instal-

Problems

Printer

Page 137 of the Philips Patient Monitor (M2540) user manual — Electromag-
Page 137 · click to enlarge
Page 137
TroubleshootingElectromagnetic Interference

Electromagnetic interference (EMI) can degrade the performance of the M2540A. Answering the

A transducer placed close to an ECG cable can increase interference. Moving the ECG cable

Answers to these questions help determine whether the problem is with the system or with the

scanning environment. Contact your local Philips service representative with a full description

  • Do two different transducers operating at the same frequency have the same problem?
  • Is the interference present if the system is moved to a different location in the facility?
  • Does the interference appear with only one transducer or with several transducers?

or other medical equipment away from the transducer can reduce interference.

following questions can help locate the source of interference:

  • Is the interference intermittent or constant?
  • Can the EMC coupling path be attenuated?

and your observations.

Electromag-

Interference

netic

Page 138 of the Philips Patient Monitor (M2540) user manual — 10 Service Procedures
Page 138 · click to enlarge
Page 138
Service ProceduresOverview

Overview This chapter includes all the service procedures that apply to the M2540A. To locate parts and

assemblies referred in the procedures, see the exploded diagrams and parts lists in “Parts” on

This manual does not include service procedures for optional peripherals. See the Philips

Ultrasound Systems Peripherals Guide for that information.

  • “Monitor Support and Front Panel” on page 160
  • “Transducer Connector Modules” on page 156
  • “System Control Panel” on page 158
  • “General Precautions” on page 139

10 Service Procedures

The chapter includes these sections:

  • “Repair Philosophy” on page 139
  • “Record Keeping” on page 141
  • “ESD Safeguards” on page 139
  • “Service Access” on page 140
  • “Monitor Twivel” on page 158
  • “Power Supply” on page 154
  • “PC Service” on page 142
  • “Monitor” on page 157
  • “E-box” on page 151
  • “Fan” on page 156

page 177.

Page 139 of the Philips Patient Monitor (M2540) user manual — Precautions
Page 139 · click to enlarge
Page 139
Service ProceduresRepair Philosophy

Follow the electrical static discharge (ESD) guidelines in this section to prevent deterioration or

Most components in the M2540A are replaced as modules. The power supply, the monitor, and

Some parts are available at reduced cost, as exchange items. See Table 13-12 on page 200 for a

  • If you use a vacuum cleaner to clean the interior of any part of the system sensitive to static

the system control panel are not serviceable, and are replaced as units. The circuit boards and

surfaces. A protective membrane under the keyboard provides some protection against spills,

  • When the system is on but not in use, leave the Power control at the lowest setting.

electricity, take the necessary ESD precautions to avoid static buildup and discharge.

Liquids, including ultrasound gel, can adversely affect the system control panel and other

but liquids may seep into electrical components under the panel and cause problems.

function cards inside the M2540A’s internal PC and e-box are replaceable in the field.

To prevent damage to the system, keep the following good practices in mind:

destruction of sensitive electronic components. General guidelines include:

  • Do not remove or install boards or cables with the system power on.
  • Lock the system’s wheels when removing or installing circuit boards.
  • Use a grounding mat for assembly work on sensitive areas.

list of exchange parts and their part numbers.

  • Turn the system off when not in use.
  • Use a properly grounded wrist strap.
  • Avoid working in carpeted areas.
  • “Cables” on page 160
  • “Cart” on page 163

Precautions

Philosophy

Safeguards

General

Repair

ESD

Page 140 of the Philips Patient Monitor (M2540) user manual — Lower Enclosure
Page 140 · click to enlarge
Page 140

You can service components mounted above the lower enclosure without opening any panels in

To service components mounted in the lower enclosure, you must open one or more panels on

Service ProceduresService Access

This section describes how to gain access to the serviceable components of the M2540A system.

the lower enclosure. For locations of the panels and components, see Figure 13-3 on page 185.

  • External peripheral components (see the Philips Ultrasound Systems Peripherals Guide)

the lower enclosure. These components include the following:

The components in the lower enclosure include the following:

  • Transducer connector modules ( page 156 )
  • Lower enclosure top panel ( page 164 )
  • System control panel ( page 158 )
  • Monitor support ( page 160 )
  • Power supply ( page 154 )
  • Lift column ( page 165 )
  • Cart wheels ( page 163 )
  • System PC ( page 142 )
  • Monitor ( page 157 )
  • Twivel ( page 158 )
  • Cables ( page 160 )
  • e-box ( page 151 )
  • Fan ( page 156 )

Lower Enclosure

Components in

Components

Above the

the Lower

Servicing

Servicing

Enclosure

Service

Access

Page 141 of the Philips Patient Monitor (M2540) user manual — Front Service
Page 141 · click to enlarge
Page 141
Service ProceduresRecord Keeping

either panel, remove the three screws along its lower edge, pull the bottom of the panel out, and

You can make notes in the RST application’s service log, to help with later service procedures. To

Doors Two doors in the rear of the lower enclosure give access to the rear of the PC, the e-box, and

Panel A panel at the right side of the system’s front gives access to the transducer connector module

Side Panels The two lower enclosure side panels are attached with screws on their lower edge. To remove

read service log entries, see “Using the Service Log” on page 125. To make an entry in the RST

the power supply. The doors are hinged near the center of the system, and are held closed by

open this panel, press down the two latches at its top, pull the top edge out, and lift the panel

mounting, the air filter, the fan, the front of the e-box, and the front of the power supply. To

quarter-turn latches near the sides of the system. To open the doors, turn the latches

service log, see “Making an Entry in the Service Log” on page 125.

lower the panel clear of the system.

counter-clockwise.

clear of the system.

system.

Front Service

Rear Service

Record

Keeping

Page 142 of the Philips Patient Monitor (M2540) user manual — Procedures
Page 142 · click to enlarge
Page 142
Service ProceduresPC Service

PC Service This section contains procedures for servicing M2540A software and the hardware components

The following procedures describe reloading and upgrading the ultrasound software, and backing

Complete the following procedure to reload or upgrade the system software. The procedure

4. Watch for the message, “Press any key to boot from CD” which displays briefly during startup. When the message appears, press the Return key.
5. When the message “Do you wish to delete all data and install XPE?” displays, press A .
1. Back up the customer’s presets. (See “Backing Up Customer Presets” on page 143 .)

the MOD drive are empty before beginning software installation.

takes about 45 minutes to complete and requires three CDs:

➤Reloading or Upgrading System Software

2. Put the Operating System CD in the CD drive.

up and restoring customer presets.

  • Ultrasound Application

System Software

3. Restart the system.
  • Operating System

of the internal PC.

  • COTS Drivers

Procedures

Software

Service

Page 143 of the Philips Patient Monitor (M2540) user manual — Backing Up and Restoring Presets
Page 143 · click to enlarge
Page 143
Service ProceduresPC Service
6. When the message, “Please insert the M2540A Ultrasound Application CD” displays, remove the Operating System CD from the CD drive and put the Ultrasound Application CD in.

installation. Do not press a key when it does. The message clears after a few moments.

Before performing any service that deletes a customer’s presets, back them up to a floppy disk.

network settings. The backup function records settings made on the DICOM Setup

window, but does not record those settings made on the Network Settings window.

7. When the message “Please insert the M2540A COTS Drivers CD” displays, remove the Ultrasound Application CD from the CD drive and put the COTS CD in. Press Y .

When the COTS software installation finishes, the system automatically restarts. If the

9. Restore the customer’s presets. (See “Restoring Customer Presets” on page 146 .)

After the procedure is complete, restore the customer’s presets from the floppy disk.

installation is successful, the monitor displays the ultrasound imaging screen.

8. When the ultrasound imaging screen appears, remove the COTS CD.
2. Insert a blank floppy disk in the PC’s floppy disk drive.

Backing Up and Restoring Presets

➤Backing Up Customer Presets

To back up a customer’s presets:

1. Start the M2540A.

Press Y.

Page 144 of the Philips Patient Monitor (M2540) user manual — Backing Up and Restoring Presets
Page 144 · click to enlarge
Page 144
Service ProceduresPC Service
3. With the ultrasound application running, press the Setup key on the system control panel. The Setup window opens.
Philips Patient Monitor (M2540) — Backing Up and Restoring Presets diagram (page 144)
Fig. 56 · Backing Up and Restoring Presets
5. Click Backup . The Backup Settings window opens.
4. Click the Floppy Disk tab.
Page 145 of the Philips Patient Monitor (M2540) user manual — Backing Up and Restoring Presets
Page 145 · click to enlarge
Page 145
Service ProceduresPC Service
7. Click Start . The backup procedure executes. When the procedure is done, the Backup Settings window closes.

system settings, the peripheral settings, and the installed options.

8. Click Close . The Setup window closes.
Philips Patient Monitor (M2540) — Backing Up and Restoring Presets diagram (page 145)
Fig. 57 · Backing Up and Restoring Presets
9. Remove the floppy disk.
6. Click Select All .
Page 146 of the Philips Patient Monitor (M2540) user manual — Restoring Customer Presets
Page 146 · click to enlarge
Page 146
Service ProceduresPC Service
2. With the ultrasound application running, press the Setup key on the system control panel. The Setup window opens (see the illustration on page 144 .)
1. Insert the floppy disk that contains the backed up presets in the PC’s floppy disk drive.
6. Select whether to restore system settings, peripheral settings, and options.
5. Click Select All , to select all of the customer’s presets for restoration.
4. Click Restore . The Restore Settings window opens.
Philips Patient Monitor (M2540) — Restoring Customer Presets diagram (page 146)
Fig. 58 · Restoring Customer Presets

➤Restoring Customer Presets

To restore a customer’s presets:

3. Click the Floppy Disk tab.
Page 147 of the Philips Patient Monitor (M2540) user manual — PC Components
Page 147 · click to enlarge
Page 147
Service ProceduresPC Service

require replacement. You can replace some PC components without removing the PC from the

replace other components, you must remove the PC from the system. (See “Removing the PC

matches the serial number of the system to that of the system that created the disk.

system, by opening the side of the PC. (See “Opening the PC Side Cover” on page 148.) To

The PC contains the following replaceable components (shown in Figure 13-1, on page 180):

PC Components None of the PC’s internal components are adjustable or repairable. Faulty PC components

8. When the restoration is complete, click Close . The Setup window closes.
  • Video capture board (part of the VCR option)
  • Sound board (part of the VCR option)

from the System” on page 150.)

9. Remove the floppy disk.
  • Graphics adapter board
  • PC motherboard
  • PC power supply
  • CD-RW drive
7. Click Start .
  • Hard drive
  • Floppy drive
  • APIO board
Page 148 of the Philips Patient Monitor (M2540) user manual — Opening the PC Side Cover
Page 148 · click to enlarge
Page 148
Service ProceduresPC Service
2. Remove the three screws from the lower edge of the lower enclosure’s left side panel, pull the bottom of the panel out, and lower the panel clear of the system.
6. If you are replacing the CD-RW drive, the Physio module, the MO drive, or the floppy disk drive, complete the following:

b. Gently pull the four plastic latches at the left side of the PC’s bezel, while pulling the left

You can gain access to components in the PC without removing the PC from the system, by

To open the PC side cover, see Figure 13-3 on page 185 to locate parts and assemblies, and

3. Remove the knurled screws from the back of the PC, and slide the cover to the rear.
5. Disconnect any cables that are connected to the component you are removing.

a. Slide the PC two cm to the left, away from the center of the system.

side of the bezel forward until the latches clear the PC case.

1. Disconnect the system power cord from the main AC supply.

wrist band, and work on a static-dissipating mat if possible.

  • Video converter board (part of the VCR option)

➤Opening the PC Side Cover

complete the following procedure:

4. Remove the PC side cover.
  • Optional Physio module

opening the PC side cover.

  • Optional MOD drive
Page 149 of the Philips Patient Monitor (M2540) user manual — Opening the PC Side Cover
Page 149 · click to enlarge
Page 149
Service ProceduresPC Service

d. Each of the replaceable parts listed in step 6 is retained by two plastic latches. To remove

the part, press the two latches in toward the part, and slide the part out of the PC case.

Reverse the procedure to reassemble the PC. Perform all RST and system checks to verify

c. Slide the bezel to the right until it can be removed.

performance.

Page 150 of the Philips Patient Monitor (M2540) user manual — Removing the PC from the System
Page 150 · click to enlarge
Page 150
Service ProceduresPC Service

To remove the PC, see Figure 13-3 on page 185 to locate parts and assemblies, and complete the

4. Loosen the screws in the angle bracket on the wall toward the center of the system. Remove the screws that secures the PC to the angle bracket.

To install the PC in the system, reverse the procedure above. Perform all RST and system checks

2. Remove the three screws from the lower edge of the lower enclosure’s left side panel, pull the bottom of the panel out, and lower the panel clear of the system.
3. Disconnect all the cables from the PC. Note their locations so you can reinstall them properly.
1. Disconnect the system power cord from the main AC supply.
5. Remove the PC through the side of the system.

➤Removing the PC from the System

to verify performance.

following procedure:

Page 151 of the Philips Patient Monitor (M2540) user manual — Ebox
Page 151 · click to enlarge
Page 151
Service ProceduresE-box

E-box You can perform most e-box service without removing the e-box from the system. To perform

grounded wrist band, and work on a static-dissipating mat if possible. See “ESD Safeguards” on

4. Lift the handles of the two latches on the e-box rear cover, and rotate them a quarter turn. Remove the cover.

service that does require removing the e-box, you must also remove the power supply (See

To replace a circuit board in the e-box, see Figure 13-2 on page 183 to locate parts and

5. If the faulty board is the BPAP board, disconnect the two cables from its face before removing it.
2. Open the door at the right rear of the lower enclosure by turning its latch counter-clockwise.
1. Disconnect the system power cord from the main AC supply.
3. Disconnect the token ring cable from the rear of the e-box.

“Removing the Power Supply and the E-box” on page 154).

assemblies, and complete the following procedure:

➤Replacing E-box Circuit Boards

6. Remove the faulty board.

disconnecting the cable.

page 139.

Page 152 of the Philips Patient Monitor (M2540) user manual — Replacing the Motherboard
Page 152 · click to enlarge
Page 152
Service ProceduresE-box

To replace the e-box motherboard, See Figure 13-2 on page 183 to locate parts and assemblies,

3. Loosen the four screws that secure each of the two septum panels in the e-box, and remove the septum panels.

grounded wrist band, and work on a static-dissipating mat when possible. See “ESD Safeguards”

Power Supply and the E-box” on page 154 and “Separating the E-box and the Power

1. Remove the BPAP board, the demodulator board, and the two TR boards from the e-box, following the steps in “Replacing E-box Circuit Boards” on page 151 .

with the e-box removed and separated from the power supply. See “Removing the

5. Tilt out the top edge of the motherboard to clear the board racks, and remove the motherboard.
2. Disconnect the power cables from the motherboard and the distribution board.
4. Remove the 15 screws that secure the motherboard to the e-box center wall.
8. Reverse step 5 through step 2 to complete the procedure.
9. Perform all RST and system checks to verify performance.

and complete the following procedure:

➤Replacing the Motherboard

7. Install the replacement board.

Supply” on page 155.

on page 139.

Page 153 of the Philips Patient Monitor (M2540) user manual — Replacing the Distribution Board
Page 153 · click to enlarge
Page 153
Service ProceduresE-box

grounded wrist band, and work on a static-dissipating mat if possible. See “ESD Safeguards” on

“Removing the Power Supply and the E-box” on page 154 and “Separating the E-box

1. Remove the BPAP board, the demodulator board, and the two TR boards from the e-box, following the steps in “Replacing E-box Circuit Boards” on page 151 .

To replace the distribution board, see Figure 13-2on page 183 to locate parts and assemblies,

3. Remove all the transducer connector modules and blank-off plates from the front of the e-box. See “Replacing Transducer Connector Modules” on page 156 .

are easier with the e-box removed and separated from the power supply. See

6. Reverse step 5 through step 1 to install the motherboard.
7. Perform all RST and system checks to verify performance.
2. Disconnect the power cable from the distribution board.

and the Power Supply” on page 155.

➤Replacing the Distribution Board

and complete the following procedure:

to the distribution board’s connectors.

page 139.

Page 154 of the Philips Patient Monitor (M2540) user manual — Power Supply
Page 154 · click to enlarge
Page 154
Service ProceduresPower Supply

Power Supply The M2540A power supply is not serviceable in the field, except for replacing the fan. The power

supply and the e-box are bolted together and must be removed from the system before they can

2. Remove the service panel at the right front of the lower enclosure by pressing down the two latches at the panel’s top.
3. Remove the two screws that secure the power supply to the angle bracket at the front of the enclosure floor.
5. Disconnect all the cables from the rear of the e-box and the power supply. Mark the cable in the top outlet. It powers the PC, and that outlet is powered whenever the system power

cable is plugged into a power source. If the PC is connected to one of the other outlets, it

To remove the power supply and the e-box from the system, see Figure 13-3 on page 185 to

4. Remove the three screws from the bottom of the lower enclosure’s right side panel, and remove the panel.

be separated. Replacement power supplies are supplied with the fan and the finger guard

4. Remove the eight screws that secure the distribution board to the e-box center wall.
5. Tilt out the top edge of the distribution board and remove the board.

locate parts and assemblies, and complete the following procedure:

1. Disconnect the system power cord from the main AC supply.
6. Reverse step 5 through step 1 to install the motherboard.
7. Perform all RST and system checks to verify performance.

➤Removing the Power Supply and the E-box

does not boot up.

installed.

Page 155 of the Philips Patient Monitor (M2540) user manual — Separating the Ebox and the Power Supply
Page 155 · click to enlarge
Page 155
Service ProceduresPower Supply
6. Remove the two screws that secure the power supply to the angle bracket at the rear of the enclosure floor.
8. Reverse step 7 through step 1 to install the power supply and the e-box in the system. Make sure the PC’s power cord is in the top outlet.
2. Disconnect the control and power cables from the BPAP board, and remove the BPAP board.
7. Lift the e-box straight up from the power supply, while feeding the cables through the hole in the bottom of the e-box.

grounded wrist band, and work on a static-dissipating mat if possible. See “ESD Safeguards” on

3. Disconnect the power cables from the motherboard and the distribution board.
7. Carefully slide the power supply and the e-box out the side of the system.
8. Reverse step 7 through step 1 to attach the e-box to the power supply.
6. Remove the four screws from the bottom of each side of the e-box.
9. Perform all RST and system checks to verify performance.
5. Remove the two screws at the bottom front of the e-box.
4. Remove the two screws at the bottom rear of the e-box.

➤Separating the E-box and the Power Supply

To separate the e-box from the power supply:

1. Remove the rear cover of the e-box.

page 139.

Page 156 of the Philips Patient Monitor (M2540) user manual — Transducer
Page 156 · click to enlarge
Page 156
Service ProceduresTransducer Connector Modules
2. Remove the service panel at the right front of the lower enclosure by pressing down the two latches at the panel’s top.

To remove the fan, see Figure 13-2 on page 183 to locate parts and assemblies, and complete the

The screws that secure the transducer connector modules are behind the service panel on the

3. Remove the screws that secure the module bezel to the e-box, remove the bezel, and pull the module out.

Fan The M2540A’s cooling fan is attached to the front of the power supply. Replacement fans are

To replace a transducer connector module, see Figure 13-2 on page 183 to locate parts and

4. Reverse step 3 through step 1 to install a transducer connector module.
1. Disconnect the system power cord from the main AC supply.
1. Disconnect the system power cord from the main AC supply.
5. Perform all RST and system checks to verify performance.

➤Replacing Transducer Connector Modules

assemblies, and complete the following procedure:

to the distribution board’s connectors.

supplied with the finger guard installed.

front of the lower enclosure.

➤Replacing the Fan

following procedure:

Transducer

Connector

Modules

Page 157 of the Philips Patient Monitor (M2540) user manual — Monitor
Page 157 · click to enlarge
Page 157
Service ProceduresMonitor
2. Remove the service panel at the right front of the lower enclosure by pressing down the two latches at the panel’s top.

field, so a monitor that does not perform properly must be replaced. See “Monitor Adjustment”

4. Use pliers or a small screwdriver to lift the heads of the four drive rivets that secure the fan to the power supply. Remove the fan.

Monitor The color monitor is a critical component of the system. The monitor is not serviceable in the

5. Align the fan for correct airflow direction (into the power supply) and connection to the power cables.

To remove the monitor from the system, see Figure 13-5 on page 193 to locate parts and

3. Press up on the latch handle under the front edge of the monitor and pull the monitor forward until it stops.
2. Disconnect the video monitor cable and the monitor power cable from the monitor.
6. Reverse step 4 through step 1 to install the fan on the power supply.
5. Reverse step 4 through step 1 to install the monitor on the system.
4. Tilt up the front of the monitor and lift the monitor off the twivel.
1. Disconnect the system power cord from the main AC supply.

on page 127 for information on adjusting the monitor display.

3. Disconnect the fan power cable from the power supply.

rivet bodies with pliers may ease installation.

assemblies, and complete the following procedure:

➤Replacing the Monitor

Page 158 of the Philips Patient Monitor (M2540) user manual — Control Panel
Page 158 · click to enlarge
Page 158

The M2540A twivel is not adjustable. The following procedure describes how to disassemble the

Service ProceduresMonitor Twivel

To reassemble the twivel assembly, perform the above procedure in reverse. When completing

3. Lift the twivel base from the cart, while carefully passing the monitor power cord and the VGA cable through the hole in the twivel base. Pass the power cord through the hole first,

step 3, pass the VGA cable’s connector through the twivel base first, before passing the power

Except for the trackball (see “Cleaning the Trackball” on page 132), the system control panel

4. Remove the four screws that secure the monitor support bezel to the monitor support.

To disassemble the twivel, see Figure 13-5 on page 193 to locate parts and assemblies, and

2. Remove the twivel shoe’s two mounting screws, and remove the twivel shoe.
1. Remove the monitor as described in “Replacing the Monitor” on page 157 .
6. Perform all RST and system checks to verify performance.

Perform all RST and system checks to verify performance.

base. Note which hole the cables go through.

contains no serviceable parts, and is replaced as a unit.

complete the following procedure:

➤Disassembling the Twivel

twivel to replace its parts.

before the video cable.

cord connector through.

Control Panel

Monitor

System

Twivel

Page 159 of the Philips Patient Monitor (M2540) user manual — Replacing the System Control Panel
Page 159 · click to enlarge
Page 159
Service ProceduresSystem Control Panel

To replace the system control panel, see Figure 13-5 on page 193 to locate parts and assemblies,

2. Remove the seven long and the nine short flathead screws that secure the control panel to the control panel surround.

the control panel standoffs. The standoffs will cut any cables underneath them when the screws

4. Disconnect the USB cable, the twisted-pair audio cables, the power cable, and the soft key ribbon cable from the control panel circuit board.
3. Push up on the bottom of the trackball, and lift the front edge of the control panel.
1. Disconnect the system power cord from the main AC supply.
6. Reverse step 5 through step 1 to install the control panel.
7. Perform all RST and system checks to verify performance.
5. Remove the system control panel from the system.

band, and work on a static-dissipating mat if possible.

➤Replacing the System Control Panel

and complete the following procedure:

are tightened.

Page 160 of the Philips Patient Monitor (M2540) user manual — Support and
Page 160 · click to enlarge
Page 160
Service ProceduresMonitor Support and Front Panel

The angular enclosure on top of the control panel surround is the monitor support. To replace

To replace the monitor support and its front panel, see Figure 13-5 on page 193 to locate parts

3. Remove the system control panel. See “Replacing the System Control Panel” on page 159 .

the mechlok cable, lift column slides, upper cable harness, or lower enclosure top panel, you

7. Remove the monitor support while feeding the VGA cable and the monitor power cable through the twivel base. Feed the monitor power cable through before the VGA cable.

Cables Most of the cables in the M2540A system pass through the lower enclosure’s central plenum.

4. From under the control panel surround, remove the 19 screws that secure the monitor support and front panel.

You may have to remove the panels from the rear of the plenum when replacing a cable.

8. Reverse step 7 through step 1 to install the monitor support and its front panel.
5. Remove the monitor support front panel by rotating its bottom edge forward.
6. Remove the twivel assembly. See “Disassembling the Twivel” on page 158 .
2. Remove the monitor. See “Replacing the Monitor” on page 157 .
1. Disconnect the system power cord from the main AC supply.

which hole the cables are in as you take the support off.

➤Replacing the Monitor Support and its Front Panel

9. Perform all RST and system checks to verify performance.

and assemblies, and complete the following procedure:

must remove the monitor support and its front panel.

Support and

Front Panel

Monitor

Page 161 of the Philips Patient Monitor (M2540) user manual — System Control
Page 161 · click to enlarge
Page 161
Service ProceduresCables
  • M2540-69060 keyboard audio line cables, which connect the PC’s sound board to the system
5. Remove the monitor support and its front panel. See “Monitor Support and Front Panel” on page 160 .

bundled together. To replace any one of these cables, you must remove the entire bundle. The

3. Remove the system control panel. See “Replacing the System Control Panel” on page 159 .
4. Remove the ties that secure the keyboard auxiliary power cable, the keyboard USB cable, and the keyboard audio line cables to the system control panel surround.

Cables that connect the lower components to the system control panel and the monitor are

  • M2540-69110 keyboard USB cable, which connects a USB port on the PC to the system
  • M2540-69050 keyboard auxiliary power cable, which connects the power supply to the
  • M2540-69180 monitor power cable, which connects the power supply to the monitor
  • M2540-69140 VGA cable, which connects the PC’s SVGA port to the monitor
2. Remove the monitor. See “Replacing the Monitor” on page 157 .
1. Disconnect the system power cord from the main AC supply.

➤Replacing the Upper Assembly Cable Bundle

  • M2540-69170 control panel ground cable

following cables comprise the bundle:

  • M2540-0457 microphone cable
  • M2540-69010 speaker cable

To replace the cable bundle:

System Control

Monitor Cables

Panel and

system control panel · control panel · control panel

Page 162 of the Philips Patient Monitor (M2540) user manual — System Control Panel and Monitor Cables
Page 162 · click to enlarge
Page 162
Service ProceduresCables
9. Disconnect the keyboard USB cable, the keyboard audio line cables, the microphone cable, and the VGA cable from the PC.
10. Feed the lower ends of the cables into the lower enclosure central plenum, while pulling on the upper end of the cable bundle.
8. Disconnect the keyboard auxiliary power cable, the monitor power cable, and the control panel ground cable from the power supply.
7. Open the two service doors at the rear of the lower enclosure by turning their latches counter-clockwise.
11. Reverse step 10 through step 1 to install the cable bundle.
12. Perform all RST and system checks to verify performance.
6. Remove the four P-clips from the speaker supports.
Page 163 of the Philips Patient Monitor (M2540) user manual — Cart
Page 163 · click to enlarge
Page 163
Service ProceduresCart

Although the front wheel assemblies on the cart are different from the rear wheels (front wheels

have swivel locks and brakes, rear wheels do not lock), the replacement procedure for all wheels

4. Carefully lift the front or the back of the cart (depending on which wheel you are replacing), and slide a support block under the front or rear edge of the lower enclosure to hold both

Each cart wheel is supplied as a caster assembly. The front caster assembly includes the integral

3. Chock both front wheels if a rear wheel is being replaced, or chock both rear wheels if a front wheel is being replaced. The chocks are a safety measure, to prevent the cart from

wheels at that end slightly off the floor. Make sure the cart is stable in this raised position.

Cart This section contains procedures for servicing the lift column, the wheels, and the lower

5. Use a 13-mm wrench to remove the four bolts that secure the wheel assembly to the bottom of the cart base, and remove the wheel.
6. Reverse step 5 through step 1 to install the new wheel assembly.
1. Disconnect the system power cord from the main AC supply.

➤Replacing Cart Wheels (Caster Assemblies)

rolling during this procedure.

To replace a caster assembly:

2. Secure the wheel locks.

brake and swivel lock.

enclosure top panel.

is the same.

Page 164 of the Philips Patient Monitor (M2540) user manual — Replacing the Top Panel of the Lower Enclosure
Page 164 · click to enlarge
Page 164
Service ProceduresCart

To replace the top panel of the lower enclosure, see Figure 13-3 on page 185 to locate parts and

5. Remove the monitor support and its front panel. See “Monitor Support and Front Panel” on page 160 .
10. Remove the e-box and the power supply. See “Removing the Power Supply and the E-box” on page 154 .
3. Remove the system control panel. See “Replacing the System Control Panel” on page 159 .
8. Remove the 11 screws that secure the system control panel surround to the column, and remove the surround.
4. Unfasten the height release handle and the cable clamp, and remove them from the cable.

You can remove and install the top panel of the lower enclosure without taking out the lift

13. Lift the top panel of the lower enclosure and tilt it to clear the lift column top flange.
7. Remove the six screws that secure the pivot latch to the top of the lift column.
9. Remove the PC. See “Removing the PC from the System” on page 150 .
12. Remove the 10 nuts that secure the top panel to the lower enclosure.
2. Remove all the peripherals from the top of the lower enclosure.
1. Disconnect the system power cord from the main AC supply.
6. Remove all the cable clamps and cable ties from the cables.

➤Replacing the Top Panel of the Lower Enclosure

assemblies, and complete the following procedure:

column or the upper cable bundle.

11. Remove the peripherals strap.

cable ends’ positions.

Page 165 of the Philips Patient Monitor (M2540) user manual — Preparation
Page 165 · click to enlarge
Page 165
Service ProceduresCart

service procedures. Because the lift column supports the weight of the upper system, you must

To prepare the system for lift column servicing, see Figure 13-4 on page 190 and Figure 13-3 on

The column is also available as an assembly, including all the components listed above. After

reassembling the system, perform all RST and system checks to assure proper performance.

The procedures for replacing the lift column components are the most complex of the field

complete the following procedure before performing any service to the column.

page 185 to locate parts and assemblies, and complete the following procedure:

14. Reverse step 13 through step 1 to install the lower enclosure top panel.
3. Remove the monitor. See “Replacing the Monitor” on page 157 .
1. Disconnect the system power cord from the main AC supply.

➤Preparing the System for Servicing the Lift Column

2. Remove all peripherals from the top of the lower enclosure.
15. Perform all RST and system checks to verify performance.

Lift Column The lift column has four replaceable components:

  • The mechlok assembly
  • Three slide assemblies
  • The mechlok cable
  • The gas spring

Preparation

the lift column.

Page 166 of the Philips Patient Monitor (M2540) user manual — Gas Spring
Page 166 · click to enlarge
Page 166
Service ProceduresCart
7. Remove the four screws on each side of the lower enclosure front center panel and the two screws at the bottom of the panel, and remove the panel.
5. Remove the e-box and power supply. See “Removing the Power Supply and the E-box” on page 154 .
9. Remove the system control panel. See “Replacing the System Control Panel” on page 159 .
10. Remove the monitor support and front panel. See “Monitor Support and Front Panel” on page 160 .

The gas spring supports the weight of the upper assembly when the system is upright. Before

4. Remove the PC. See “Removing the PC from the System” on page 150 .
11. Rotate the control panel surround to face the rear of the system.
6. Remove the cable hook on the front of the lower enclosure.

working on the gas spring, you must relieve the weight.

12. Carefully lay the system down on its rear side.
8. Extend the lift column to its highest position.

➤Replacing the Gas Spring

and damage to the system.

To replace the gas spring:

Gas Spring

Page 167 of the Philips Patient Monitor (M2540) user manual — Mechlok and Cable
Page 167 · click to enlarge
Page 167
Service ProceduresCart
2. Remove the three screws that secure the mechlok handle to the control panel surround, and unclip the handle from the cable end.

The mechlok assembly locks the lift column at any height in the column’s range. The mechlok is

3. Swing the top of the gas spring out, and unscrew the gas spring shaft from the pivot at the bottom of the cart.

surround. You can replace the cable without completely removing the mechlok, and you can

controlled by a cable connected to a handle under the front of the system control panel

2. Remove the two screws that secure the upper gas spring bracket to the column.
4. Unclip and remove the retaining spring from the ball-cup end of the gas spring.
5. Pull the gas spring’s cup end off the ball stud on the gas spring upper bracket.

replace the mechlok mechanism without removing the cable from the system.

7. Perform all RST and system checks to verify performance.
6. Reverse step 5 through step 1 to install the gas spring.
1. Complete “Preparation” on page 165 .

➤Replacing the Mechlok and Cable

1. Complete “Preparation” on page 165 .

To remove the mechlok and cable:

and damage to the system.

Mechlok and Cable

Page 168 of the Philips Patient Monitor (M2540) user manual — Mechlok and Cable
Page 168 · click to enlarge
Page 168
Service ProceduresCart
4. From under the bottom of the system, remove the two screws that secure the gas spring and mechlok lower bracket to the floor of the lower enclosure.
9. If you are replacing the mechlok, remove the retaining clip from the pivot pin in the gas spring and the mechlok lower bracket, and remove the pin.
11. After installation and righting of the system, adjust the position of the cable clamp under the control panel surround to provide proper locking and unlocking.
6. Grasp the ring on the mechlok locking pin, and pull the pin out of the column. Remove the mechlok and gas spring.
8. If you are replacing the cable, pull the upper end of the cable up through the control panel surround and down through the lift column.
3. Remove the two screws that secure the cable clamp to the control panel surround, and unclip the clamp from the cable outer sheath.
5. Remove the two screws that secure the upper gas spring bracket to the column.
7. Slide the cable locking clip off the tab on the mechlok to remove the cable.
10. Reverse step 9 through step 1 to install the mechlok and cable.
12. Perform all RST and system checks to verify performance.
Page 169 of the Philips Patient Monitor (M2540) user manual — Slides
Page 169 · click to enlarge
Page 169
Service ProceduresCart

When replacing slides, it is not necessary to remove the mechlok cable, or to disconnect the

5. Push the lower ends of the upper cable bundle into the plenum, to prevent their snagging in the plenum openings.
7. Remove the four nylock nuts that secure the lift column surround plate to the top panel of the lower enclosure.
3. Remove the two screws at the bottom of the rear center panel, near the door hinges, and remove the panel and the doors.
4. Remove the gas spring and mechlok assembly. See “Replacing the Mechlok and Cable” on page 167 .

Each of the three lift column slide assemblies consists of an inner slide mounted on the lift

2. Remove the I/O panel and the anchor panel for the peripheral strap.

column and an outer slide mounted in the lower enclosure plenum.

8. Carefully pull the lift column out of the lower enclosure.

mechlok or the gas spring from their lower bracket.

1. Complete “Preparation” on page 165 .

To replace the lift column slides:

6. Stand the system upright.

➤Replacing the Slides

and damage to the system.

Slides

Page 170 of the Philips Patient Monitor (M2540) user manual — Slides
Page 170 · click to enlarge
Page 170
Service ProceduresCart
10. Remove the four screws that secure each inner slide to the lift column, and remove the inner slides.
12. Run the column through its entire range of motion. If it binds, loosen all the outer slide mounting nuts, run the column through its entire range of motion, and then tighten the nuts.
9. Remove the four nylock nuts that secure each outer slide to the lower enclosure plenum, and remove the outer slides from the lower enclosure.
11. Reverse step 10 through step 1 to install the lift column slides.

ball carriers.

Page 171 of the Philips Patient Monitor (M2540) user manual — 11 Cabling
Page 171 · click to enlarge
Page 171

Cabling:

The following table lists the cables in the M2540A. The cables may also be listed in assembly

Description Part Number Qty Reference

Cart to external AC power cable M2540-69120 1 PC (DP) power cable 8120-2572 1 Table 13-5

Internal LAN cable M2540-69130 1 Table 13-5

Composite video RCA cable M2540-69140 1 Trigger miniphone cable M2540-69150 1 Table 13-5

Printer USB cable M2540-69100 2 Peripheral power cable (IEC) 8120-2572 3 Table 13-5

I/O panel foot switch cable M2540-69160 1 Table 13-5

Token ring cable M2540-69000 1 Table 13-5

Ground cable M2540-69170 1 System cable harness M2540-69220 1 Table 13-5

Monitor power cable M2540-69180 1 Monitor video cable M2540-69040 1 Speaker cable M2540-69010 1 Keyboard USB cable M2540-69110 1 Keyboard auxiliary power cable M2540-69050 1

parts lists shown in the “Reference” column of this table.

11 Cabling

Table 11-1 Cables

Page 172 of the Philips Patient Monitor (M2540) user manual — 11 Cabling
Page 172 · click to enlarge
Page 172

Cabling:

Description Part Number Qty Reference

Keyboard line audio cable M2540-69060 1 Power cords

UK power cord M2540-60900 1 Australia power cord M2540-60901 1 Europe power cord M2540-60902 1 US, Canada, and Japan power cord M2540-60903 1 Switzerland power cord M2540-60906 1 Denmark power cord M2540-60912 1 South Africa power cord M2540-60917 1 Israel power cord M2540-60919 1 Argentina power cord M2540-60920 1 PRC power cord M2540-60922 1 Physio option cables

Physio option USB cable M2540-69020 1 Physio option trigger cable M2540-69030 1 3-Lead AAMI grabber set M1603A

3-Lead AAMI trunk cable M1500A

3 Lead AAMI mini-clip set M1609A

3-lead IEC trunk cable M1510A

Table 11-1 Cables (Continued)

Page 173 of the Philips Patient Monitor (M2540) user manual — 11 Cabling
Page 173 · click to enlarge
Page 173

Cabling:

Description Part Number Qty Reference

VCR option video in/out cable M2540-69070 2 VCR option audio in/out cable M2540-69200 2 VCR option serial cable M2540-69090 1 Optional accessory cables

External LAN cable M2540-69080

USB cable 10-foot M2540-69110

Table 11-1 Cables (Continued)

VCR option cables

Page 174 of the Philips Patient Monitor (M2540) user manual — Hard Drive
Page 174 · click to enlarge
Page 174
ConfigurationDisk Drive Settings

information on the correct settings. The hard disk drive is the only device on the primary IDE,

The jumper positions and switch settings of the drives in the M2540A depend on the

Jumpers Figure 12-1shows typical hard disk drive jumper positions. Examine the drive itself for

Philips Patient Monitor (M2540) — 12 Configuration diagram (page 174)
Fig. 59 · 12 Configuration

and is usually set to MASTER or SINGLE.

Figure 12-1 Hard Disk Drive Jumper Settings

manufacturer of the drives.

12 Configuration

Disk Drive

Hard Drive

Settings

Page 175 of the Philips Patient Monitor (M2540) user manual — MO Drive
Page 175 · click to enlarge
Page 175
ConfigurationVCR Switch Settings

Jumpers Figure 12-2 shows typical CD-RW drive jumper positions. Set the jumper to the MASTER

Jumpers Set the MO drive jumpers to SLAVE on the secondary IDE.

The VCR’s switches must be set to the following positions:

Figure 12-2 CD-RW Drive Jumper Position

  • Audio Out right switch: Norm

position on the secondary IDE.

  • Audio Out left switch: Mix
Philips Patient Monitor (M2540) — CD Drive Jumpers diagram (page 175)
Fig. 60 · CD Drive Jumpers
  • Mode Lock: Off
  • Input: S-Video
  • S-VHS: Auto
  • Menu: Off

VCR Switch

MO Drive

CD Drive

Settings

Page 176 of the Philips Patient Monitor (M2540) user manual — Printer Dip
Page 176 · click to enlarge
Page 176
ConfigurationSony B/W Printer Dip Switch Settings

On the Sony black-and-white printer, dip switch number 5 should be up. All the other dip

switches should be down.

Printer Dip

Sony B/W

Switch

Settings

Page 177 of the Philips Patient Monitor (M2540) user manual — Ordering
Page 177 · click to enlarge
Page 177
PartsOverview

Overview This section contains lists of the M2540A’s parts, supplies, and accessories available for use with

Address all parts orders or inquiries to your local Philips Sales and Service office. Always specify

the correct part number or exchange number for the assembly or component being ordered, as

well as the serial and revision numbers of the system in which the assembly is being installed.

the system, and information on ordering parts, accessories, and supplies.

  • “Transducer Connector Modules” on page 179
  • “Supplies and Accessories” on page 202
  • “e-box and Power Supply” on page 183
  • “Exchange Part Numbers” on page 200
  • “Lift Column Assembly” on page 190
  • “Physio Option Parts” on page 199
  • “Optional Peripherals” on page 201
  • “VCR Option Parts” on page 198
  • “Major Assemblies” on page 178
  • “Lower Assembly” on page 185
  • “Upper Assembly” on page 193
  • “Ordering Parts” on page 177
  • “Internal PC” on page 180
  • “Top Shelf” on page 196
  • “Software” on page 198

13 Parts

Ordering

Parts

Page 178 of the Philips Patient Monitor (M2540) user manual — System Serial
Page 178 · click to enlarge
Page 178
PartsMajor Assemblies

New Parts Be sure to specify the correct sub-assembly or component part number, as listed throughout this

Exchange Parts Some circuit boards, transducers, and OEM peripherals are available as exchange items (repaired

rear of the cart. The ID label also identifies installed system options. Use the system Help file to

or refurbished), at reduced cost. Exchange item part numbers are different from original equip-

When ordering any new or exchange part, be sure to include the system serial number and the

software revision prefix (for example, B.0). The serial number is on the system ID label, on the

The following table lists the major assemblies of the M2540A, and the drawings that show them.

find software revision numbers, or run the RST application, and under the Info tab, click Sys-

To learn about transducer parts, please see the M2540A Transducer Guide.

PC Figure 13-1 on page 180

Power supply Figure 13-2 on page 183

e-box Figure 13-2 on page 183

Upper assembly Figure 13-5 on page 193

Lower assembly Figure 13-3 on page 185

Lift column Figure 13-4 on page 190

ment part numbers. See “Exchange Part Numbers” on page 200

Description Drawing and Page

tem. (See “System Information” on page 109.)

Table 13-1 Major Assemblies

section.

System Serial

Revision Level

Number and

Assemblies

Major

Page 179 of the Philips Patient Monitor (M2540) user manual — Transducer
Page 179 · click to enlarge
Page 179
PartsTransducer Connector Modules

The following table lists the available transducer connector modules and their part numbers.

Description Part Number Qty

Cannon transducer connector module M2540-62100 1 Advance Vision transducer connector module M2540-62000 1 Pencil transducer connector module M2540-62200 1 Blank module M2540-62010 1

Table 13-2 Transducer Connector Modules

Transducer

Connector

Modules

Page 180 of the Philips Patient Monitor (M2540) user manual — Internal PC
Page 180 · click to enlarge
Page 180
PartsInternal PC

Internal PC The following figure shows the replaceable components of the M2540A’s PC. Numbers refer to

Philips Patient Monitor (M2540) — Internal PC diagram (page 180)
Fig. 61 · Internal PC

Figure 13-1 Internal PC Components

Table 13-3, on page 181.

Page 181 of the Philips Patient Monitor (M2540) user manual — Internal PC
Page 181 · click to enlarge
Page 181
PartsInternal PC

The following table lists the replaceable components of the internal PC shown in Figure 13-1.

Numbers in the # column are the item numbers on the drawing. A dash (—) in the # column

# Description Part Number Qty Reference

PC M2540-67000 1 Figure 13-3

1 APIO board M2540-60030 1 2 PC motherboard M2540-67110 1 3 CPU M2540-67140 1 4 Hard disk drive M2540-67100 1 5 CD-RW drive M2540-67130 1 6 Floppy disk drive M2540-67120 1 7 PC power supply M2540-67150 1 8 PA 810 enclosure M2540-67170 1 9 PC bezel M2540-67210 1 — Ribbon Cable kit M2540-67160 1 — 256 MB memory module M2540-67180 1 — 512 MB memory module M2540-67190 1 — Drive bay blank cover kit M2540-67200 1 — Screw, 6-32 x 5/8, panhead 2360-0123 3

Table 13-3 Internal PC Components

means the part is not numbered on the drawing.

Page 182 of the Philips Patient Monitor (M2540) user manual — Internal PC
Page 182 · click to enlarge
Page 182
PartsInternal PC

# Description Part Number Qty Reference

Optional physio module Table 13-11

10 Physio module M2540-62300 1 11 Physio USB cable M2540-69020 1 Table 11-1

12 Physio trigger cable M2540-69030 1 Table 11-1

13 MOD drive M2540-80100 1 — MOD adapter assembly M2540-80400 1 PC components for VCR option — — Table 13-9

14 Video converter board M2540-80070 1 Table 13-9

15 Video capture board M2540-80080 1 Table 13-9

16 Sound card M3150-60020 1 Table 13-9

Table 13-3 Internal PC Components (Continued)

Optional magneto optical drive — —

Page 183 of the Philips Patient Monitor (M2540) user manual — Power Supply
Page 183 · click to enlarge
Page 183
Partse-box and Power Supply

The following figure shows the system power supply and the components of the e-box listed in

Philips Patient Monitor (M2540) — e-box and Power Supply diagram (page 183)
Fig. 62 · e-box and Power Supply

Table 13-4. The power supply contains no replaceable parts.

Figure 13-2 e-box and Power Supply Components

Power Supply

e-box and

Page 184 of the Philips Patient Monitor (M2540) user manual — e-box and Power Supply
Page 184 · click to enlarge
Page 184

Figure 13-2. Numbers in the # column are the item numbers on the drawing. A dash (—) in the

Partse-box and Power Supply

The following table lists the components of the power supply and e-box assembly shown in

2 Cooling fan M2540-66010 1 3 e-box assembly M2540-64000 1 Does not include circuit boards

1 Power supply M2540-66000 1 Exchange part, see Table 13-12

4 Motherboard M2540-60000 1 5 Distribution board M2540-60010 1 6 TR board M2540-60040 2 Exchange part, see Table 13-12

7 Demodulator board M2540-60050 1 Exchange part, see Table 13-12

8 BPAP board M2540-60060 1 Exchange part, see Table 13-12

Table 13-4 Power Supply and e-box Assembly Components

# Description Part Number Qty Reference

# column means the part is not numbered on the drawing.

9 Chassis septum M2540-00040 2 10 Card guide 4040-2551 8 11 RFI shielding kit M2540-86040 1 12 M3 x 6mm screw 0515-0430 27 13 M4 x 10-mm screw M2540-80322 16 14 e-box door assembly M2540-64010 1

Page 185 of the Philips Patient Monitor (M2540) user manual — Assembly
Page 185 · click to enlarge
Page 185
PartsLower Assembly

The following figure shows the components of the lower assembly listed in Table 13-5.

Philips Patient Monitor (M2540) — Lower Assembly diagram (page 185)
Fig. 63 · Lower Assembly

Figure 13-3 Lower Assembly

Assembly

Lower

Page 186 of the Philips Patient Monitor (M2540) user manual — Lower Assembly
Page 186 · click to enlarge
Page 186
PartsLower Assembly

The following table lists the components of the lower assembly shown in Figure 13-3. Numbers

in the # column are the item numbers on the drawing. A dash (—) in the # column means the

2 Air filter M2540-80050 1 3 e-box and power supply — 1 Bolted together, see Figure 13-2

8 Side panel, right M2540-00400 1 9 Side panel, left assembly M2540-65190 1 10 Rear panel assembly M2540-65170 1 Includes rear doors

# Description Part Number Qty Reference

5 Control top — 1 Figure 13-5

6 VCR — 1 Table 13-13

7 Printer — 1 Table 13-13

Cart assembly M2540-65000 1 1 PC M2540-67000 1 Table 13-3

4 Column assembly M2540-65100 1 Table 13-6

11 Cart top M2540-00420 1 12 Center column, front M2540-00370 1 13 Latch, rear door 1390-1326 2 14 Front panel M2540-00440 1 15 Top front trim M2540-00590 1 16 Upper I/O panel M2540-00660 1

Table 13-5 Lower Assembly

part is not numbered on the drawing.

Page 187 of the Philips Patient Monitor (M2540) user manual — Lower Assembly
Page 187 · click to enlarge
Page 187
PartsLower Assembly

# Description Part Number Qty Reference

17 Lower I/O panel M2540-00670 1 LAN bulkhead connector 1252-6178 1 Grounding stud 1251-5964 1 I/O panel trigger cable M2540-69150 1 Table 11-1

I/O panel foot switch cable M2540-69160 1 Table 11-1

Internal LAN cable M2540-69130 1 Table 11-1

Composite video cable M2540-69140 1 Table 11-1

18 Cart front panel bracket M2540-00620 1 19 Power supply bracket M2540-00530 2 20 PC bracket M2540-00600 1 21 Transducer cable hook M2540-40690 4 22 Cable hook, rear M2540-40680 1 23 External power plug and cable M2540-69120 1 24 Peripheral strap 77922-80290 1 25 Strap bracket kit M2540-86020 1 — Color printer tray kit M2540-86010 1 26 VCR tray M2540-00810 1 — Printer spacer M2540-40750 4 — Nylon bushing, 2-inch 0040-0951 1 — VCR option video in/out cable M2540-69070 2 Table 13-9

— VCR option audio in/out cable M2409-69200 2 Table 13-9

Table 13-5 Lower Assembly (Continued)

Page 188 of the Philips Patient Monitor (M2540) user manual — Lower Assembly
Page 188 · click to enlarge
Page 188
PartsLower Assembly

# Description Part Number Qty Reference

— VCR option serial cable M2540-69090 1 Table 13-9

— Printer USB cable M2540-69100 2 — Peripheral power cable (IEC) 8120-2572 3 Table 11-1

Table 13-5 Lower Assembly (Continued)

— M3 x 12-mm screw 0515-0664 10 28 M4 nylock nut 0535-1029 12 — Foam strip M2540-80120 4 29 Center cart assembly M2540-65140 1 30 Plain swivel caster M2540-80240 2 31 Locking swivel caster M2540-80250 2 32 M8 x 16-mm hex locking bolt 0515-2466 16 — M4 x 10-mm screw with

— Token ring cable M2540-69000 1 — Ground cable M2540-69170 1 — Power cord M2540-40700 1 27 M4 x 10-mm Phillips locking

— M6 lock nut 0535-1030 4 — M5 x 16-mm screw 0515-0671 1 — M5 plastic hex nut 0535-0090 1 — Nameplate BASIC M2540-83010 1

washer M2540-80322 39 · screw 0515-2859 4

Page 189 of the Philips Patient Monitor (M2540) user manual — Lower Assembly
Page 189 · click to enlarge
Page 189
PartsLower Assembly

# Description Part Number Qty Reference

Table 13-5 Lower Assembly (Continued)

— Nameplate Performance M2540-83020 1 — System Serial Overlay Label M2540-83030

— Performance label strip M2540-83040 1 — Nameplate CV M2540-83050 1 — Nameplate CV performance M2540-83060 1

Page 190 of the Philips Patient Monitor (M2540) user manual — Lift Column
Page 190 · click to enlarge
Page 190
PartsLift Column Assembly

The following figure shows the components of the lift column assembly listed in Table 13-6.

Philips Patient Monitor (M2540) — Lift Column Assembly diagram (page 190)
Fig. 64 · Lift Column Assembly

Figure 13-4 Lift Column

Lift Column

Assembly

9 15 12

Page 191 of the Philips Patient Monitor (M2540) user manual — M2540-65180
Page 191 · click to enlarge
Page 191
PartsLift Column Assembly

The following table lists the components of the lift column assembly shown in Figure 13-4. Num-

bers in the # column are the item numbers on the drawing. A dash (—) in the # column means

6, 7, and strut eyelet 4 Mechlok bracket — 1 Part of

1 Column assembly M2540-65100 1 2 Slide assembly M2540-65030 3 3 Gas spring bracket M2540-00310 1 — Mechlok bracket hardware kit M2540-65180 — Includes items 4, 5,

M2540-65180

M2540-65180

M2540-65180

M2540-65180

# Description Part Number Qty Reference

5 Grommet, 5/16-inch ID — 4 Part of

6 Flat washer, 5/16-inch — 4 Part of

7 Clevis pin — 2 Part of

8 Gas strut assembly M2540-80020 1 9 Column cover M2540-00430 1 10 Mechlok M2540-80030 1 11 Slide bracket, front M2540-00320 1 — Mechlock cable M2540-80035 1

Table 13-6 Lift Column Components

the part is not numbered on the drawing.

Page 192 of the Philips Patient Monitor (M2540) user manual — Lift Column Assembly
Page 192 · click to enlarge
Page 192
PartsLift Column Assembly

# Description Part Number Qty Reference

Table 13-6 Lift Column Components (Continued)

— Friction ring M2540-80010 2 — Belleville washer 3050-1403 8 12 Detent ring M2540-00270 1 13 Cable protector cap M2540-40710 1 14 Quick release pin and ring 1480-1141 1 15 Pivot mount M2540-00230 1 16 M4 x 6-mm panhead screw 0515-0684 12 — M4 x 10-mm screw with washer M2540-80322 12 — M4 x 10-mm screw with washer,

— M4 x 10-mm screw with washer, light gray M2540-80325 2 17 M4 lock nut 0535-1029 12

mushroom M2540-80320 10

Page 193 of the Philips Patient Monitor (M2540) user manual — Assembly
Page 193 · click to enlarge
Page 193
PartsUpper Assembly

The following figure shows the components of the upper assembly listed in Table 13-7.

Philips Patient Monitor (M2540) — Upper Assembly diagram (page 193)
Fig. 65 · Upper Assembly

Figure 13-5 Upper Assembly

Assembly

Upper

Page 194 of the Philips Patient Monitor (M2540) user manual — Upper Assembly
Page 194 · click to enlarge
Page 194
PartsUpper Assembly

column are the item numbers on the drawing. A dash (—) in the # column means the part is not

Table 13-7 lists the components of the upper assembly shown in Figure 13-5. Numbers in the #

1 Top shelf assembly 1 Figure 13-6 on page 196

# Description Part Number Qty Reference

6 Cable stop bracket M2540-00550 1 7 M4 x 16-mm flathead screw 0515-4620 12 8 M4 x 25-mm flathead screw 0515-4622 25 9 M4 x 25-mm panhead screw 0515-2860 2 10 Transducer cable hook M2540-40690 3 — Twivel assembly

— FIMI monitor, mushroom M2540-66500 1 2 System control panel (keyboard) M2540-40100 1 3 Height release handle M2540-40560 1 4 Monitor support M2540-00210 1 5 Monitor support front panel

11 Twivel ring M2540-40740 1 12 Twivel shoe M2540-40730 1 13 Monitor mount (twivel base) M2540-40630 1 14 Twivel hat M2540-40610 1

Table 13-7 Upper Assembly Components

numbered on the drawing.

assembly M2540-65120 1

Page 195 of the Philips Patient Monitor (M2540) user manual — Upper Assembly
Page 195 · click to enlarge
Page 195
PartsUpper Assembly

22 M4 nylock nut 0535-1029 2 — M4 x 10-mm screw with washer M2540-80322 2 23 Trackball assembly M2540-80200 1 Includes items 24 and 25

— Pivot mount assembly 1 — Pivot mount M2540-00230 1 15 Pivot knob M2540-40550 1 16 Latch cam M2540-40590 1 17 Pivot latch assembly M2540-65150 1 18 Latch bearing M2540-40600 1 19 M4 x 6-mm Phillips screw 0515-2859 3 Used with items 20, 21

# Description Part Number Qty Reference

Table 13-7 Upper Assembly Components (Continued)

24 Trackball M2540-80210 1 25 Trackball ring M2540-80220 1 — Control panel replacement knob kit M2540-86070 —

— Rotary encoder M2540-40210 7 — Cable tie 1400-0401 12

20 Delrin flat washer 3050-2184 3 21 Precision flat washer

0.171-inch ID 3050-2186 3

Page 196 of the Philips Patient Monitor (M2540) user manual — Top Shelf
Page 196 · click to enlarge
Page 196
PartsTop Shelf
Philips Patient Monitor (M2540) — Top Shelf diagram (page 196)
Fig. 66 · Top Shelf

Top Shelf The following figure shows the components of the upper assembly listed in Table 13-8.

Figure 13-6 Top Shelf

Page 197 of the Philips Patient Monitor (M2540) user manual — Top Shelf
Page 197 · click to enlarge
Page 197
PartsTop Shelf

The following table lists the parts shown in Figure 13-6. Numbers in the # column are the item

numbers on the drawing. A dash (—) in the # column means the part is not numbered on the

1 Bottom sheet M2540-00160 1 2 Keyboard surround M2540-40500 1 3 Top sheet M2540-00150 1 4 Speaker support, left M2540-00190 1 5 Speaker support, right M2540-00200 1 6 Front panel retainer clip M2540-00880 4 7 Speaker M2540-80230 2 8 P-clip, 1-inch 2 Part of M2540-86050

9 P-clip, 5/8-inch 1 Part of M2540-86050

10 P-clip, 3/16-inch 2 Part of M2540-86050

# Description Part Number Qty Reference

11 M4 x 10-mm Phillips screw 0515-2859 4 12 M3 x 6-mm Phillips screw 0515-0430 8 — Speaker foam M2540-80110 2

Table 13-8 Top Shelf Components

drawing.

Page 198 of the Philips Patient Monitor (M2540) user manual — VCR Option
Page 198 · click to enlarge
Page 198
PartsVCR Option Parts

The following table lists the parts of the VCR option. For part numbers of available VCRs, see

Software The following table lists the part numbers for the M2540A’s software CDs.

Description Part Number Qty

Ultrasound application CD M2540-81000-01

COTS CD M2540-82000-01

Operating system CD M2540-82500-01

Video converter board M2540-80070 1 Video capture board M2540-80080 1 Sound card M3150-60020 1 Video in/out cable M2540-69070 2 Audio in/out cable M2450-69200 1 Serial cable M2540-69090 1

Table 13-10 M2540A Software CDs

Table 13-9 VCR Option Parts

“Optional Peripherals” on page 201.

VCR Option

Parts

Description Part Number

Page 199 of the Philips Patient Monitor (M2540) user manual — Physio Option
Page 199 · click to enlarge
Page 199
PartsPhysio Option Parts

Description Part Number Qty Reference

Physio module (ECG module) M2540-62300 1 Table 13-3

Physio USB cable M2540-69020 1 Table 11-1

Physio trigger cable M2540-69030 1 Table 11-1

ECG electrodes 77020-62180 Pediatric electrodes, disposable 77020-62190

The following table lists the parts of the physio option.

3-Lead AAMI grabber set M1603A

3-Lead AAMI trunk cable M1500A

3 Lead AAMI mini-clip set M1609A

3-lead IEC trunk cable M1510A

3-lead IEC grabber set M1613A

3-lead IEC mini-clip set M1619A

Table 13-11 Physio Option Parts

Physio Option

Parts

Page 200 of the Philips Patient Monitor (M2540) user manual — Exchange Part
Page 200 · click to enlarge
Page 200
PartsExchange Part Numbers

Some parts are available at reduced cost, as exchange items. The following table lists the part

Description Part Number Exchange Part

FIMI monitor M2540-66500 M2540-66508

System power supply M2540-66000 M2540-66008

TR board M2540-60040 M2540-60048

Demodulator board M2540-60050 M2540-60058

BPAP board M2540-60060 M2540-60068

NTSC VCR 77616A 77616-68000 PAL VCR 77626A 77626-68000 Black-and-white printer UPD-895 (USB) M2540-66700 M2540-66708

Color printer UPD-21MD (USB) M2540-66800 M2540-66808

S4 ultraband sector transducer 21330A 21330-68000 S8 ultraband sector transducer 21350A 21350-68000 S12 transducer 21380A 21380-68000 c3540 biopsy capable transducer 21321A 21320-68000 EC6509 endocavity transducer 21336A 21336-68000 L7535 linear transducer 21359A 21359-68000

Table 13-12 Exchange Part Numbers

numbers for exchange parts and assemblies.

Exchange Part

Numbers

Keyboard M2540-40100 M2540-40108 · Number

Page 201 of the Philips Patient Monitor (M2540) user manual — Peripherals
Page 201 · click to enlarge
Page 201
PartsOptional Peripherals

NTSC VCR 77616A Exchange part, see Table 13-12

PAL VCR 77626A Exchange part, see Table 13-12

Black-and-white printer UPD-895 (USB) M2540-66700 Exchange part, see Table 13-12

Color printer UPD-21MD (USB) M2540-66800 Exchange part, see Table 13-12

Description Part Number Exchange Part

The following table lists the part numbers of peripheral options available for the M2540A.

L5035 linear transducer 21360A 21360-68000 TEE (Omniplane II) transducer 21369-60000 21369-68000 L1038A small parts transducer 21376A 21376-68000 15-6L intraoperative transducer 21390A 21390-68000 PA 4-2 sector transducer 21422A 21422-68000 CA 5-2 transducer 21425A 21425-68000

Description Part Number Note

Table 13-12 Exchange Part Numbers

Color plain-paper printer HP 940C

Black-and-white plain paper laser printer HP1200

Table 13-13 Optional Peripherals

Peripherals

Optional

Number · MOD drive M2540-80100

Page 202 of the Philips Patient Monitor (M2540) user manual — Supplies and
Page 202 · click to enlarge
Page 202
PartsSupplies and Accessories

The following table lists supplies and accessories available from Philips for use with the M2540A.

To order supplies and accessories, see “Ordering Parts” on page 177. To learn about transducer

Microphone 9164-0719 Foot switch M2540-80280

External LAN cable M2540-69080

USB cable 10-foot M2540-69110

Table 13-14 Supplies and Accessories

accessories, please see the M2540A Transducer Guide.

compliance.

Supplies and

Accessories

Show 4 more labels / short text on this page
Description Part Number · Shipping package M2540-85000 · MO disk M2540-80101 · Floppy disk 9164-0361
Page 203 of the Philips Patient Monitor (M2540) user manual — Supplies and Accessories
Page 203 · click to enlarge
Page 203
PartsSupplies and Accessories

black-and-white Sony digital graphic printer UP-D895 SNY110HG

black-and-white Sony digital graphic printer UP-D895MD SNY110HD

Sony UPP-110S standard paper 5 roll box SNY110S

Aquasonic gel, US, 12 8.5-oz. bottles 40483A Aquasonic gel, US, 1 5-liter-bottle refill 40483B Ultrasound gel, Europe, 12 bottles 40404A Ultrasound gel, Europe, 1 5-liter-bottle refill 40404B

Table 13-14 Supplies and Accessories (Continued)

Sony UPC 21 L color printing pack for the Sony digital color

Sony UPP-110 HD high density paper printing pack for the

Sony UPP-110 HG glossy paper printing pack for the

Show 4 more labels / short text on this page
Description Part Number · printer UP-D21MD SNY21L · Ultrasound Gel · Paper
Page 204 of the Philips Patient Monitor (M2540) user manual — Supplies and Accessories
Page 204 · click to enlarge
Page 204
PartsSupplies and Accessories

Biopsy kit C3540 Civco 645-057

Biopsy kit PA4-2 Civco 645-069

Biopsy kit CA5-2 Civco 645-067

Biopsy kit L1038 Civco 645-059

Biopsy kits L7535 and L5035 Civco 645-039

Biopsy kit E6509 Civco 610-588

Foam electrodes, pregelled, 100 packs 40420A Pediatric electrodes, disposable 13951C ECG Lead Sets

3-lead set, grabber, AAMI M1603A

3 lead trunk cable, AAMI M1500A

3-lead set, snap, AAMI M1605A

3-lead set, mini-clip, AAMI M1609A

3 lead trunk cable, IEC M1510A

3-lead set, grabber, IEC M1613A

3-lead set, snap, IEC M1615A

3-lead set, mini-clip, IEC M1619A

Table 13-14 Supplies and Accessories (Continued)

Description Part Number · ECG electrodes · Biopsy kits

Page 205 of the Philips Patient Monitor (M2540) user manual — 14 Transducers
Page 205 · click to enlarge
Page 205

Transducers:

exchange items. See Table 13-12 on page 200 for a list of exchange part numbers.

PA 4-2 sector 21422A 2-4 MHz Advance Vision

CA 5-2 21425A 2-5 MHz Advance Vision

Description Part Number Frequency Connector

S4 ultraband sector1 21330A Cannon

S8 ultraband sector 21350A Cannon

c3540 biopsy capable1 21321A 3.5 MHz Cannon

EC6509 endocavity 21336A Cannon

L7535 linear 21359A 7.5 MHz Cannon

L5035 linear 21360A 5 MHz Cannon

TEE (Omniplane II) 21369-60000 6 MHz Cannon

L1038A small parts 21376A Cannon

15-6L intraoperative 21390A Cannon

Pencil 21221-60062 1.9 MHz Pencil

1. This transducer is compatible with the M2540A, but is only sold separately.

Table 14-1 lists the transducers you can use with the M2540A.

14 Transducers

S12 21380A Cannon · Table 14-1 Transducers

Page 206 of the Philips Patient Monitor (M2540) user manual — 15 Glossary
Page 206 · click to enlarge
Page 206

Glossary:

2D mode A mode of operation where the display represents reflected ultrasound information as points of

Angio mode This mode presents two-dimensional, real-time Doppler or time shift information superimposed

Color mode This mode presents two-dimensional, real-time Doppler or time shift information superimposed

information, regardless of delay. The detected frequency shifts can produce audio signals, or they

colored pixel information represents the velocities of all the blood flows, or the flow velocities,

as colored pixels on a real-time, gray scale anatomic cross-sectional image. The colored pixel

CW Doppler mode In the Continuous Wave Doppler mode, the transducer array uses separate transmitting and

information represents the amplitudes of the selected blood flows within the angio window.

as colored pixels on a real-time, gray scale anatomic cross-sectional image. Flow directions

toward and away from the transducer are presented as different colors on the display. The

Control Any mechanism (softkey, touch key, push button, toggle, rotary knob, etc.) that changes a

receiving piezoelectric elements. It processes the entire continuum of reflected signal

light of varying intensity arrayed linearly, concordant with the arrival time.

Default preset A preset created at the factory and shipped with the ultrasound system.

parameter that, in turn, affects the displayed image.

can display graphically in a spectral trace.

Control setting The current setting of an active control.

Archive A depository for the final reports.

Custom preset A preset created by the user.

within the color window.

15 Glossary

Page 207 of the Philips Patient Monitor (M2540) user manual — 15 Glossary
Page 207 · click to enlarge
Page 207

Glossary:

Image An ultrasound view that is either actively displayed or archived. An image can be represented by

Display All the information visible on the system's monitor, including ultrasound images, measurements,

a single frame, a continuous loop, a 3D volume set, a Doppler spectral trace, Doppler audio, an

images, network archive of images, all measurements and calculations, image annotations, and

ECG Electrocardiogram. The system displays a live ECG waveform if one is connected and activated.

reverse video (light-on-dark, or dark-on-light, whichever is the reverse of the normal display).

Live mode Any unfrozen system mode that shows acoustic data in real-time. Live mode includes scanning

Keys on the keyboard are highlighted by lighting an adjacent LED. Softkeys are highlighted by

Live imaging mode Any unfrozen system mode that includes a two-dimensional image being created in real-time

Documentation Everything produced as a result of an ultrasound exam: prints, film, VCR tape, M/O disk of

Highlight An indication that a control is active. Text on a CRT screen is highlighted by displaying it in

Imaging mode 2D, Color, Angio, Mmode, PW Doppler, CW Doppler, 3D, etc.

(that is, not retrieved from VCR tape, Cine Loop memory, etc.).

with a non-imaging transducer. Compare to Live Imaging mode.

the sonographer's or physician's exam-related comments.

tion system The computer databases and networks of an institution.

changing the value associated with them.

system state, patient information, etc.

Institution A hospital, lab, clinic, or office.

ECG waveform, etc.

Hospital informa-

Page 208 of the Philips Patient Monitor (M2540) user manual — (PW)
Page 208 · click to enlarge
Page 208

Glossary:

Mmode Method of image processing in which the display is recorded as a spot of light for each detected

A mode that uses pulsed ultrasound. The detected frequency shifts can produce audio signals, or

Sonographer A non-physician who has been trained to perform ultrasound studies. The sonographer performs

sound pulse returning from a moving target. Produces a single-dimensional time/motion display

Scanning window An area of the 2D image pertaining to one of three specific modes: 2D, Color, or Zoom. The

the patient study directly on the patient and captures results for later review by a reviewing

Setup parameters System settings that can be configured to accommodate the user's requirements and

Peripherals Output devices such as VCR, MOD, hard-copy printers, network interface, etc.

windows pertaining to these areas are called the 2D, Flow, and Zoom windows

Preset A group of control settings that optimizes the system for a particular exam.

System text Text showing settings of parameters having to do with system operation.

System settings Refers to the values of a set of controls of the ultrasound system.

Select To indicate a choice, as in choosing an item from a menu.

MOD Magneto-optical disk; a type of removable data storage.

Option A piece of system functionality that a user can purchase.

Modality work list A list of patient names and associated studies to occur.

they can display graphically as a spectral trace.

of detected anatomical structures.

preferences.

physician.

(PW)

Doppler mode · Pulsed Wave

Page 209 of the Philips Patient Monitor (M2540) user manual — 15 Glossary
Page 209 · click to enlarge
Page 209

Glossary:

Transducer icon A graphic symbol that represents the ultrasonic transducer. This symbol indicates the scan plane

and flow imaging using pulse-echo sound waves that are greater than 20 kHz (above the range of

Transducer The element of the ultrasound system that the sonographer places on the patient. Sometimes

called a probe. The transducer emits high-frequency sound waves into the patient and receives

their echoes. The ultrasound system controls the transmitting and receiving of the ultrasound

Ultrasound system Any system that can perform diagnostic sonography, which is medical cross-sectional anatomic

processes the received echoes to construct the displayed images, also in accordance with the

waves according to imaging parameters set by the sonographer. The ultrasound system

by placing it in the proper position and orientation to a body mark icon.

imaging parameters set by the sonographer.

human hearing).

Page 210 of the Philips Patient Monitor (M2540) user manual — Table of contents
Page 210 · click to enlarge
Page 210

Table of contents

1 About This Manualp.1About This Manualp.1Audiencep.2Formatp.3Conventionsp.3Service Manual Questions or Commentsp.3Customer Assistancep.41 General Informationp.6Product Overviewp.6System Descriptionp.7Cartp.7Cart Wheelsp.7PCp.7System Monitorp.8e-boxp.8System Control Panelp.8Physio Modulep.9Power Supplyp.9Optional Peripherals (VCR, Printers, and Image Devices)p.9Preset Functionsp.10Keyboard Equivalenciesp.11Safety Specificationsp.12Safety Limitsp.12Regulatory Compliancep.122 Specificationsp.13Physical Dimensionsp.14Electrical Specificationsp.19Monitorp.19Transducer Specificationsp.20Connection and Communication Specificationsp.21System I/O Portsp.21PC Portsp.21I/O Panel Portsp.23Physio Port Specificationsp.23Audio/Video Specificationsp.23Audiop.24External Videop.24Video Output Specificationsp.253 Safetyp.26Transmit Power (Acoustical)p.27Acoustic Exposurep.27AIUM/NEMA Output Display Standardp.28Automatic Index Selectionp.29When MI is Displayed With the Normal Settingp.29When TIS is Displayed With Normal Settingp.29When TIC is Displayed With Normal Settingp.30Explosive Hazardsp.30Electrical Warningsp.31Peripheral Connectionsp.31Glutaraldehyde Exposurep.32Moving the Systemp.32Before Moving the Systemp.32When Moving the Systemp.33Moving on Ramps or Uneven Surfacesp.33System Tiltingp.33Transporting the System in a Vehiclep.34Electromagnetic Compatibilityp.35Avoiding Electromagnetic Interferencep.35Restrictions for Usep.36Immunity Level Test Resultsp.36Electrostatic Dischargep.36Electrosurgical Unitsp.374 Theory of Operationp.38Overviewp.38Internal PCp.39Standard PC Componentsp.39PC Portsp.39VCR-Option Components in the PCp.40Optional Drive-Bay Componentsp.40TR Boardsp.40BPAP Boardp.42Omni TEE Circuitryp.42Demodulator Boardp.43APIO Boardp.43Physio Module Optionp.43Primary Functionsp.44Connectorsp.44System Motherboardp.44Distribution Board and Connector Modulesp.44Transducer Connector Module Typesp.45I/O Panelp.45System Control Panelp.45Power Supplyp.46Functional Block Diagramsp.47System Functional Block Diagramp.47e-box and Power Supply Functional Block Diagramp.49Token Ring Cablep.50Physio Functional Block Diagramp.51Transducer Safety Testing: Test Setup and Theoryp.52Theory of the Transducer Leakage Current Testp.545 Installationp.59Overviewp.59Assemble the Systemp.59Power Cord and Monitor Installationp.59Installing the Power Cord and Monitorp.59Peripherals Installationp.59Installing a VCRp.60Installing a Sony Color Printerp.60Installing a Sony Black-and-White Printerp.61System Startupp.62Powering On the Systemp.62Monitor Signal Voltagep.62Setting the Monitor Signal Input Voltagep.62Printer Preparationp.64Preparing the Sony Black-and-White Printerp.64Preparing the Sony Color Printerp.64System Configurationp.64Localizationp.65Setting the Localization Parametersp.65Options Installationp.71Installing Optionsp.71DICOM Network Configurationp.73Changing the System’s Network Parametersp.73Using the Field Service Options Window for Advanced Network Settingsp.74Adding Server Informationp.77Assigning Servers to DICOM Rolesp.78Configuring Autodeletep.79Configuring Loggingp.81System Hardware Configurationp.83Assigning Record Buttonsp.83Assigning Option Buttonsp.84VCR configurationp.85Configuring the System for PAL Formatp.85Setting the Monitor Color Temperaturep.87Configuring for printersp.87Customer Trainingp.886 Performance Testsp.89Overviewp.89Electrical Safety Testsp.89Chassis to Ground Resistance Testp.90Chassis to Ground Resistance Test Procedurep.91Ground Wire Leakage Current Testp.92Ground Wire Leakage Current Test Procedurep.93ECG Lead Leakage Current Test (Source)p.94ECG Lead Leakage Current Test Procedurep.95ECG Lead Isolation Leakage Current Testp.96ECG Lead Isolation Leakage Current Test Procedurep.97Transducer Testsp.98Transducer Leakage Current Test (Source)p.99Transducer Leakage Current Test Procedurep.100Transducer Isolation Leakage Current Testp.102Transducer Isolation Leakage Current Test Procedurep.103PC Diagnosticsp.105IBM Drive Fitness Testp.105Creating a DFT Bootable Floppy Diskp.105Testing PC Functions Using the Drive Fitness Testp.105Checkit Prop.106Testing PC Functions Using Checkit Prop.106Resident Self Testsp.107Availabilityp.107Logging Onp.107User Interfacep.108Information Typesp.109System Informationp.109Hardware Informationp.110Testsp.111Available Testsp.111Testing Optionsp.112Test Resultsp.113Service Logp.113Launching the RST Applicationp.113Closing the RST Applicationp.114Running a Basic Testp.114Running an Extended Testp.115Test Modulesp.117Running an Interactive Testp.118Viewing Error Logsp.119Ultrasound Application Logp.119Clearing the Ultrasound Application Logp.120RST Test Logp.121RST Log Codesp.122Clearing the RST Test Logp.122Test Result Detailsp.123Viewing Test Detailsp.123System Logp.124Clearing the System Logp.124Using the Service Logp.125Making an Entry in the Service Logp.125Modifying an Entry in the Service Logp.1267 Adjustmentsp.127Monitor Adjustmentp.127OSD Main Menup.127Opening the OSD Main Menup.127Navigating the OSD Menusp.128Adjusting VGA Vertical Sizep.128Adjust the VGA Vertical Sizep.128Changing the Boot Sequencep.130Change the Boot Sequencep.1308 Preventive Maintenancep.131Cleaning the Air Filterp.131Cleaning the Trackballp.132Cleaning the e-boxp.132Degaussing the Monitorp.1339 Troubleshootingp.134PC Won’t Bootp.134PC Boots, but Monitor is Blankp.135PC Boots, But Peripherals Don’t Workp.135Printer Problemsp.136HP940 Startupp.136Printer Installation Problemsp.136Electromagnetic Interferencep.13710 Service Proceduresp.138Overviewp.138Repair Philosophyp.139ESD Safeguardsp.139General Precautionsp.139Service Accessp.140Servicing Components Above the Lower Enclosurep.140Servicing Components in the Lower Enclosurep.140Rear Service Doorsp.141Front Service Panelp.141Side Panelsp.141Record Keepingp.141PC Servicep.142Software Service Proceduresp.142System Softwarep.142Reloading or Upgrading System Softwarep.142Backing Up and Restoring Presetsp.143Backing Up Customer Presetsp.143Restoring Customer Presetsp.146PC Componentsp.147Opening the PC Side Coverp.148Removing the PC from the Systemp.150Eboxp.151Replacing Ebox Circuit Boardsp.151Replacing the Motherboardp.152Replacing the Distribution Boardp.153Power Supplyp.154Removing the Power Supply and the Eboxp.154Separating the Ebox and the Power Supplyp.155Transducer Connector Modulesp.156Replacing Transducer Connector Modulesp.156Fanp.156Replacing the Fanp.156Monitorp.157Replacing the Monitorp.157Monitor Twivelp.158Disassembling the Twivelp.158System Control Panelp.158Replacing the System Control Panelp.159Monitor Support and Front Panelp.160Replacing the Monitor Support and its Front Panelp.160Cablesp.160System Control Panel and Monitor Cablesp.161Replacing the Upper Assembly Cable Bundlep.161Cartp.163Replacing Cart Wheels (Caster Assemblies)p.163Replacing the Top Panel of the Lower Enclosurep.164Lift Columnp.165Preparationp.165Preparing the System for Servicing the Lift Columnp.165Gas Springp.166Replacing the Gas Springp.166Mechlok and Cablep.167Replacing the Mechlok and Cablep.167Slidesp.169Replacing the Slidesp.16911 Cablingp.17112 Configurationp.174Disk Drive Settingsp.174Hard Drive Jumpersp.174CD Drive Jumpersp.175MO Drive Jumpersp.175VCR Switch Settingsp.175Sony B/W Printer Dip Switch Settingsp.17613 Partsp.177Overviewp.177Ordering Partsp.177New Partsp.178Exchange Partsp.178System Serial Number and Revision Levelp.178Major Assembliesp.178Transducer Connector Modulesp.179Internal PCp.180e-box and Power Supplyp.183Lower Assemblyp.185Lift Column Assemblyp.190Upper Assemblyp.193Top Shelfp.196VCR Option Partsp.198Softwarep.198Physio Option Partsp.199Exchange Part Numbersp.200Optional Peripheralsp.201Supplies and Accessoriesp.20214 Transducersp.20515 Glossaryp.206
Page 211 of the Philips Patient Monitor (M2540) user manual — Table of contents
Page 211 · click to enlarge
Page 211

Table of contents

1 About This Manualp.1About This Manualp.1Audiencep.2Formatp.3Conventionsp.3Service Manual Questions or Commentsp.3Customer Assistancep.41 General Informationp.6Product Overviewp.6System Descriptionp.7Cartp.7Cart Wheelsp.7PCp.7System Monitorp.8e-boxp.8System Control Panelp.8Physio Modulep.9Power Supplyp.9Optional Peripherals (VCR, Printers, and Image Devices)p.9Preset Functionsp.10Keyboard Equivalenciesp.11Safety Specificationsp.12Safety Limitsp.12Regulatory Compliancep.122 Specificationsp.13Physical Dimensionsp.14Electrical Specificationsp.19Monitorp.19Transducer Specificationsp.20Connection and Communication Specificationsp.21System I/O Portsp.21PC Portsp.21I/O Panel Portsp.23Physio Port Specificationsp.23Audio/Video Specificationsp.23Audiop.24External Videop.24Video Output Specificationsp.253 Safetyp.26Transmit Power (Acoustical)p.27Acoustic Exposurep.27AIUM/NEMA Output Display Standardp.28Automatic Index Selectionp.29When MI is Displayed With the Normal Settingp.29When TIS is Displayed With Normal Settingp.29When TIC is Displayed With Normal Settingp.30Explosive Hazardsp.30Electrical Warningsp.31Peripheral Connectionsp.31Glutaraldehyde Exposurep.32Moving the Systemp.32Before Moving the Systemp.32When Moving the Systemp.33Moving on Ramps or Uneven Surfacesp.33System Tiltingp.33Transporting the System in a Vehiclep.34Electromagnetic Compatibilityp.35Avoiding Electromagnetic Interferencep.35Restrictions for Usep.36Immunity Level Test Resultsp.36Electrostatic Dischargep.36Electrosurgical Unitsp.374 Theory of Operationp.38Overviewp.38Internal PCp.39Standard PC Componentsp.39PC Portsp.39VCR-Option Components in the PCp.40Optional Drive-Bay Componentsp.40TR Boardsp.40BPAP Boardp.42Omni TEE Circuitryp.42Demodulator Boardp.43APIO Boardp.43Physio Module Optionp.43Primary Functionsp.44Connectorsp.44System Motherboardp.44Distribution Board and Connector Modulesp.44Transducer Connector Module Typesp.45I/O Panelp.45System Control Panelp.45Power Supplyp.46Functional Block Diagramsp.47System Functional Block Diagramp.47e-box and Power Supply Functional Block Diagramp.49Token Ring Cablep.50Physio Functional Block Diagramp.51Transducer Safety Testing: Test Setup and Theoryp.52Theory of the Transducer Leakage Current Testp.545 Installationp.59Overviewp.59Assemble the Systemp.59Power Cord and Monitor Installationp.59Installing the Power Cord and Monitorp.59Peripherals Installationp.59Installing a VCRp.60Installing a Sony Color Printerp.60Installing a Sony Black-and-White Printerp.61System Startupp.62Powering On the Systemp.62Monitor Signal Voltagep.62Setting the Monitor Signal Input Voltagep.62Printer Preparationp.64Preparing the Sony Black-and-White Printerp.64Preparing the Sony Color Printerp.64System Configurationp.64Localizationp.65Setting the Localization Parametersp.65Options Installationp.71Installing Optionsp.71DICOM Network Configurationp.73Changing the System’s Network Parametersp.73Using the Field Service Options Window for Advanced Network Settingsp.74Adding Server Informationp.77Assigning Servers to DICOM Rolesp.78Configuring Autodeletep.79Configuring Loggingp.81System Hardware Configurationp.83Assigning Record Buttonsp.83Assigning Option Buttonsp.84VCR configurationp.85Configuring the System for PAL Formatp.85Setting the Monitor Color Temperaturep.87Configuring for printersp.87Customer Trainingp.886 Performance Testsp.89Overviewp.89Electrical Safety Testsp.89Chassis to Ground Resistance Testp.90Chassis to Ground Resistance Test Procedurep.91Ground Wire Leakage Current Testp.92Ground Wire Leakage Current Test Procedurep.93ECG Lead Leakage Current Test (Source)p.94ECG Lead Leakage Current Test Procedurep.95ECG Lead Isolation Leakage Current Testp.96ECG Lead Isolation Leakage Current Test Procedurep.97Transducer Testsp.98Transducer Leakage Current Test (Source)p.99Transducer Leakage Current Test Procedurep.100Transducer Isolation Leakage Current Testp.102Transducer Isolation Leakage Current Test Procedurep.103PC Diagnosticsp.105IBM Drive Fitness Testp.105Creating a DFT Bootable Floppy Diskp.105Testing PC Functions Using the Drive Fitness Testp.105Checkit Prop.106Testing PC Functions Using Checkit Prop.106Resident Self Testsp.107Availabilityp.107Logging Onp.107User Interfacep.108Information Typesp.109System Informationp.109Hardware Informationp.110Testsp.111Available Testsp.111Testing Optionsp.112Test Resultsp.113Service Logp.113Launching the RST Applicationp.113Closing the RST Applicationp.114Running a Basic Testp.114Running an Extended Testp.115Test Modulesp.117Running an Interactive Testp.118Viewing Error Logsp.119Ultrasound Application Logp.119Clearing the Ultrasound Application Logp.120RST Test Logp.121RST Log Codesp.122Clearing the RST Test Logp.122Test Result Detailsp.123Viewing Test Detailsp.123System Logp.124Clearing the System Logp.124Using the Service Logp.125Making an Entry in the Service Logp.125Modifying an Entry in the Service Logp.1267 Adjustmentsp.127Monitor Adjustmentp.127OSD Main Menup.127Opening the OSD Main Menup.127Navigating the OSD Menusp.128Adjusting VGA Vertical Sizep.128Adjust the VGA Vertical Sizep.128Changing the Boot Sequencep.130Change the Boot Sequencep.1308 Preventive Maintenancep.131Cleaning the Air Filterp.131Cleaning the Trackballp.132Cleaning the e-boxp.132Degaussing the Monitorp.1339 Troubleshootingp.134PC Won’t Bootp.134PC Boots, but Monitor is Blankp.135PC Boots, But Peripherals Don’t Workp.135Printer Problemsp.136HP940 Startupp.136Printer Installation Problemsp.136Electromagnetic Interferencep.13710 Service Proceduresp.138Overviewp.138Repair Philosophyp.139ESD Safeguardsp.139General Precautionsp.139Service Accessp.140Servicing Components Above the Lower Enclosurep.140Servicing Components in the Lower Enclosurep.140Rear Service Doorsp.141Front Service Panelp.141Side Panelsp.141Record Keepingp.141PC Servicep.142Software Service Proceduresp.142System Softwarep.142Reloading or Upgrading System Softwarep.142Backing Up and Restoring Presetsp.143Backing Up Customer Presetsp.143Restoring Customer Presetsp.146PC Componentsp.147Opening the PC Side Coverp.148Removing the PC from the Systemp.150Eboxp.151Replacing Ebox Circuit Boardsp.151Replacing the Motherboardp.152Replacing the Distribution Boardp.153Power Supplyp.154Removing the Power Supply and the Eboxp.154Separating the Ebox and the Power Supplyp.155Transducer Connector Modulesp.156Replacing Transducer Connector Modulesp.156Fanp.156Replacing the Fanp.156Monitorp.157Replacing the Monitorp.157Monitor Twivelp.158Disassembling the Twivelp.158System Control Panelp.158Replacing the System Control Panelp.159Monitor Support and Front Panelp.160Replacing the Monitor Support and its Front Panelp.160Cablesp.160System Control Panel and Monitor Cablesp.161Replacing the Upper Assembly Cable Bundlep.161Cartp.163Replacing Cart Wheels (Caster Assemblies)p.163Replacing the Top Panel of the Lower Enclosurep.164Lift Columnp.165Preparationp.165Preparing the System for Servicing the Lift Columnp.165Gas Springp.166Replacing the Gas Springp.166Mechlok and Cablep.167Replacing the Mechlok and Cablep.167Slidesp.169Replacing the Slidesp.16911 Cablingp.17112 Configurationp.174Disk Drive Settingsp.174Hard Drive Jumpersp.174CD Drive Jumpersp.175MO Drive Jumpersp.175VCR Switch Settingsp.175Sony B/W Printer Dip Switch Settingsp.17613 Partsp.177Overviewp.177Ordering Partsp.177New Partsp.178Exchange Partsp.178System Serial Number and Revision Levelp.178Major Assembliesp.178Transducer Connector Modulesp.179Internal PCp.180e-box and Power Supplyp.183Lower Assemblyp.185Lift Column Assemblyp.190Upper Assemblyp.193Top Shelfp.196VCR Option Partsp.198Softwarep.198Physio Option Partsp.199Exchange Part Numbersp.200Optional Peripheralsp.201Supplies and Accessoriesp.20214 Transducersp.20515 Glossaryp.206
Page 212 of the Philips Patient Monitor (M2540) user manual — Table of contents
Page 212 · click to enlarge
Page 212

Table of contents

1 About This Manualp.1About This Manualp.1Audiencep.2Formatp.3Conventionsp.3Service Manual Questions or Commentsp.3Customer Assistancep.41 General Informationp.6Product Overviewp.6System Descriptionp.7Cartp.7Cart Wheelsp.7PCp.7System Monitorp.8e-boxp.8System Control Panelp.8Physio Modulep.9Power Supplyp.9Optional Peripherals (VCR, Printers, and Image Devices)p.9Preset Functionsp.10Keyboard Equivalenciesp.11Safety Specificationsp.12Safety Limitsp.12Regulatory Compliancep.122 Specificationsp.13Physical Dimensionsp.14Electrical Specificationsp.19Monitorp.19Transducer Specificationsp.20Connection and Communication Specificationsp.21System I/O Portsp.21PC Portsp.21I/O Panel Portsp.23Physio Port Specificationsp.23Audio/Video Specificationsp.23Audiop.24External Videop.24Video Output Specificationsp.253 Safetyp.26Transmit Power (Acoustical)p.27Acoustic Exposurep.27AIUM/NEMA Output Display Standardp.28Automatic Index Selectionp.29When MI is Displayed With the Normal Settingp.29When TIS is Displayed With Normal Settingp.29When TIC is Displayed With Normal Settingp.30Explosive Hazardsp.30Electrical Warningsp.31Peripheral Connectionsp.31Glutaraldehyde Exposurep.32Moving the Systemp.32Before Moving the Systemp.32When Moving the Systemp.33Moving on Ramps or Uneven Surfacesp.33System Tiltingp.33Transporting the System in a Vehiclep.34Electromagnetic Compatibilityp.35Avoiding Electromagnetic Interferencep.35Restrictions for Usep.36Immunity Level Test Resultsp.36Electrostatic Dischargep.36Electrosurgical Unitsp.374 Theory of Operationp.38Overviewp.38Internal PCp.39Standard PC Componentsp.39PC Portsp.39VCR-Option Components in the PCp.40Optional Drive-Bay Componentsp.40TR Boardsp.40BPAP Boardp.42Omni TEE Circuitryp.42Demodulator Boardp.43APIO Boardp.43Physio Module Optionp.43Primary Functionsp.44Connectorsp.44System Motherboardp.44Distribution Board and Connector Modulesp.44Transducer Connector Module Typesp.45I/O Panelp.45System Control Panelp.45Power Supplyp.46Functional Block Diagramsp.47System Functional Block Diagramp.47e-box and Power Supply Functional Block Diagramp.49Token Ring Cablep.50Physio Functional Block Diagramp.51Transducer Safety Testing: Test Setup and Theoryp.52Theory of the Transducer Leakage Current Testp.545 Installationp.59Overviewp.59Assemble the Systemp.59Power Cord and Monitor Installationp.59Installing the Power Cord and Monitorp.59Peripherals Installationp.59Installing a VCRp.60Installing a Sony Color Printerp.60Installing a Sony Black-and-White Printerp.61System Startupp.62Powering On the Systemp.62Monitor Signal Voltagep.62Setting the Monitor Signal Input Voltagep.62Printer Preparationp.64Preparing the Sony Black-and-White Printerp.64Preparing the Sony Color Printerp.64System Configurationp.64Localizationp.65Setting the Localization Parametersp.65Options Installationp.71Installing Optionsp.71DICOM Network Configurationp.73Changing the System’s Network Parametersp.73Using the Field Service Options Window for Advanced Network Settingsp.74Adding Server Informationp.77Assigning Servers to DICOM Rolesp.78Configuring Autodeletep.79Configuring Loggingp.81System Hardware Configurationp.83Assigning Record Buttonsp.83Assigning Option Buttonsp.84VCR configurationp.85Configuring the System for PAL Formatp.85Setting the Monitor Color Temperaturep.87Configuring for printersp.87Customer Trainingp.886 Performance Testsp.89Overviewp.89Electrical Safety Testsp.89Chassis to Ground Resistance Testp.90Chassis to Ground Resistance Test Procedurep.91Ground Wire Leakage Current Testp.92Ground Wire Leakage Current Test Procedurep.93ECG Lead Leakage Current Test (Source)p.94ECG Lead Leakage Current Test Procedurep.95ECG Lead Isolation Leakage Current Testp.96ECG Lead Isolation Leakage Current Test Procedurep.97Transducer Testsp.98Transducer Leakage Current Test (Source)p.99Transducer Leakage Current Test Procedurep.100Transducer Isolation Leakage Current Testp.102Transducer Isolation Leakage Current Test Procedurep.103PC Diagnosticsp.105IBM Drive Fitness Testp.105Creating a DFT Bootable Floppy Diskp.105Testing PC Functions Using the Drive Fitness Testp.105Checkit Prop.106Testing PC Functions Using Checkit Prop.106Resident Self Testsp.107Availabilityp.107Logging Onp.107User Interfacep.108Information Typesp.109System Informationp.109Hardware Informationp.110Testsp.111Available Testsp.111Testing Optionsp.112Test Resultsp.113Service Logp.113Launching the RST Applicationp.113Closing the RST Applicationp.114Running a Basic Testp.114Running an Extended Testp.115Test Modulesp.117Running an Interactive Testp.118Viewing Error Logsp.119Ultrasound Application Logp.119Clearing the Ultrasound Application Logp.120RST Test Logp.121RST Log Codesp.122Clearing the RST Test Logp.122Test Result Detailsp.123Viewing Test Detailsp.123System Logp.124Clearing the System Logp.124Using the Service Logp.125Making an Entry in the Service Logp.125Modifying an Entry in the Service Logp.1267 Adjustmentsp.127Monitor Adjustmentp.127OSD Main Menup.127Opening the OSD Main Menup.127Navigating the OSD Menusp.128Adjusting VGA Vertical Sizep.128Adjust the VGA Vertical Sizep.128Changing the Boot Sequencep.130Change the Boot Sequencep.1308 Preventive Maintenancep.131Cleaning the Air Filterp.131Cleaning the Trackballp.132Cleaning the e-boxp.132Degaussing the Monitorp.1339 Troubleshootingp.134PC Won’t Bootp.134PC Boots, but Monitor is Blankp.135PC Boots, But Peripherals Don’t Workp.135Printer Problemsp.136HP940 Startupp.136Printer Installation Problemsp.136Electromagnetic Interferencep.13710 Service Proceduresp.138Overviewp.138Repair Philosophyp.139ESD Safeguardsp.139General Precautionsp.139Service Accessp.140Servicing Components Above the Lower Enclosurep.140Servicing Components in the Lower Enclosurep.140Rear Service Doorsp.141Front Service Panelp.141Side Panelsp.141Record Keepingp.141PC Servicep.142Software Service Proceduresp.142System Softwarep.142Reloading or Upgrading System Softwarep.142Backing Up and Restoring Presetsp.143Backing Up Customer Presetsp.143Restoring Customer Presetsp.146PC Componentsp.147Opening the PC Side Coverp.148Removing the PC from the Systemp.150Eboxp.151Replacing Ebox Circuit Boardsp.151Replacing the Motherboardp.152Replacing the Distribution Boardp.153Power Supplyp.154Removing the Power Supply and the Eboxp.154Separating the Ebox and the Power Supplyp.155Transducer Connector Modulesp.156Replacing Transducer Connector Modulesp.156Fanp.156Replacing the Fanp.156Monitorp.157Replacing the Monitorp.157Monitor Twivelp.158Disassembling the Twivelp.158System Control Panelp.158Replacing the System Control Panelp.159Monitor Support and Front Panelp.160Replacing the Monitor Support and its Front Panelp.160Cablesp.160System Control Panel and Monitor Cablesp.161Replacing the Upper Assembly Cable Bundlep.161Cartp.163Replacing Cart Wheels (Caster Assemblies)p.163Replacing the Top Panel of the Lower Enclosurep.164Lift Columnp.165Preparationp.165Preparing the System for Servicing the Lift Columnp.165Gas Springp.166Replacing the Gas Springp.166Mechlok and Cablep.167Replacing the Mechlok and Cablep.167Slidesp.169Replacing the Slidesp.16911 Cablingp.17112 Configurationp.174Disk Drive Settingsp.174Hard Drive Jumpersp.174CD Drive Jumpersp.175MO Drive Jumpersp.175VCR Switch Settingsp.175Sony B/W Printer Dip Switch Settingsp.17613 Partsp.177Overviewp.177Ordering Partsp.177New Partsp.178Exchange Partsp.178System Serial Number and Revision Levelp.178Major Assembliesp.178Transducer Connector Modulesp.179Internal PCp.180e-box and Power Supplyp.183Lower Assemblyp.185Lift Column Assemblyp.190Upper Assemblyp.193Top Shelfp.196VCR Option Partsp.198Softwarep.198Physio Option Partsp.199Exchange Part Numbersp.200Optional Peripheralsp.201Supplies and Accessoriesp.20214 Transducersp.20515 Glossaryp.206
Page 213 of the Philips Patient Monitor (M2540) user manual — Table of contents
Page 213 · click to enlarge
Page 213

Table of contents

1 About This Manualp.1About This Manualp.1Audiencep.2Formatp.3Conventionsp.3Service Manual Questions or Commentsp.3Customer Assistancep.41 General Informationp.6Product Overviewp.6System Descriptionp.7Cartp.7Cart Wheelsp.7PCp.7System Monitorp.8e-boxp.8System Control Panelp.8Physio Modulep.9Power Supplyp.9Optional Peripherals (VCR, Printers, and Image Devices)p.9Preset Functionsp.10Keyboard Equivalenciesp.11Safety Specificationsp.12Safety Limitsp.12Regulatory Compliancep.122 Specificationsp.13Physical Dimensionsp.14Electrical Specificationsp.19Monitorp.19Transducer Specificationsp.20Connection and Communication Specificationsp.21System I/O Portsp.21PC Portsp.21I/O Panel Portsp.23Physio Port Specificationsp.23Audio/Video Specificationsp.23Audiop.24External Videop.24Video Output Specificationsp.253 Safetyp.26Transmit Power (Acoustical)p.27Acoustic Exposurep.27AIUM/NEMA Output Display Standardp.28Automatic Index Selectionp.29When MI is Displayed With the Normal Settingp.29When TIS is Displayed With Normal Settingp.29When TIC is Displayed With Normal Settingp.30Explosive Hazardsp.30Electrical Warningsp.31Peripheral Connectionsp.31Glutaraldehyde Exposurep.32Moving the Systemp.32Before Moving the Systemp.32When Moving the Systemp.33Moving on Ramps or Uneven Surfacesp.33System Tiltingp.33Transporting the System in a Vehiclep.34Electromagnetic Compatibilityp.35Avoiding Electromagnetic Interferencep.35Restrictions for Usep.36Immunity Level Test Resultsp.36Electrostatic Dischargep.36Electrosurgical Unitsp.374 Theory of Operationp.38Overviewp.38Internal PCp.39Standard PC Componentsp.39PC Portsp.39VCR-Option Components in the PCp.40Optional Drive-Bay Componentsp.40TR Boardsp.40BPAP Boardp.42Omni TEE Circuitryp.42Demodulator Boardp.43APIO Boardp.43Physio Module Optionp.43Primary Functionsp.44Connectorsp.44System Motherboardp.44Distribution Board and Connector Modulesp.44Transducer Connector Module Typesp.45I/O Panelp.45System Control Panelp.45Power Supplyp.46Functional Block Diagramsp.47System Functional Block Diagramp.47e-box and Power Supply Functional Block Diagramp.49Token Ring Cablep.50Physio Functional Block Diagramp.51Transducer Safety Testing: Test Setup and Theoryp.52Theory of the Transducer Leakage Current Testp.545 Installationp.59Overviewp.59Assemble the Systemp.59Power Cord and Monitor Installationp.59Installing the Power Cord and Monitorp.59Peripherals Installationp.59Installing a VCRp.60Installing a Sony Color Printerp.60Installing a Sony Black-and-White Printerp.61System Startupp.62Powering On the Systemp.62Monitor Signal Voltagep.62Setting the Monitor Signal Input Voltagep.62Printer Preparationp.64Preparing the Sony Black-and-White Printerp.64Preparing the Sony Color Printerp.64System Configurationp.64Localizationp.65Setting the Localization Parametersp.65Options Installationp.71Installing Optionsp.71DICOM Network Configurationp.73Changing the System’s Network Parametersp.73Using the Field Service Options Window for Advanced Network Settingsp.74Adding Server Informationp.77Assigning Servers to DICOM Rolesp.78Configuring Autodeletep.79Configuring Loggingp.81System Hardware Configurationp.83Assigning Record Buttonsp.83Assigning Option Buttonsp.84VCR configurationp.85Configuring the System for PAL Formatp.85Setting the Monitor Color Temperaturep.87Configuring for printersp.87Customer Trainingp.886 Performance Testsp.89Overviewp.89Electrical Safety Testsp.89Chassis to Ground Resistance Testp.90Chassis to Ground Resistance Test Procedurep.91Ground Wire Leakage Current Testp.92Ground Wire Leakage Current Test Procedurep.93ECG Lead Leakage Current Test (Source)p.94ECG Lead Leakage Current Test Procedurep.95ECG Lead Isolation Leakage Current Testp.96ECG Lead Isolation Leakage Current Test Procedurep.97Transducer Testsp.98Transducer Leakage Current Test (Source)p.99Transducer Leakage Current Test Procedurep.100Transducer Isolation Leakage Current Testp.102Transducer Isolation Leakage Current Test Procedurep.103PC Diagnosticsp.105IBM Drive Fitness Testp.105Creating a DFT Bootable Floppy Diskp.105Testing PC Functions Using the Drive Fitness Testp.105Checkit Prop.106Testing PC Functions Using Checkit Prop.106Resident Self Testsp.107Availabilityp.107Logging Onp.107User Interfacep.108Information Typesp.109System Informationp.109Hardware Informationp.110Testsp.111Available Testsp.111Testing Optionsp.112Test Resultsp.113Service Logp.113Launching the RST Applicationp.113Closing the RST Applicationp.114Running a Basic Testp.114Running an Extended Testp.115Test Modulesp.117Running an Interactive Testp.118Viewing Error Logsp.119Ultrasound Application Logp.119Clearing the Ultrasound Application Logp.120RST Test Logp.121RST Log Codesp.122Clearing the RST Test Logp.122Test Result Detailsp.123Viewing Test Detailsp.123System Logp.124Clearing the System Logp.124Using the Service Logp.125Making an Entry in the Service Logp.125Modifying an Entry in the Service Logp.1267 Adjustmentsp.127Monitor Adjustmentp.127OSD Main Menup.127Opening the OSD Main Menup.127Navigating the OSD Menusp.128Adjusting VGA Vertical Sizep.128Adjust the VGA Vertical Sizep.128Changing the Boot Sequencep.130Change the Boot Sequencep.1308 Preventive Maintenancep.131Cleaning the Air Filterp.131Cleaning the Trackballp.132Cleaning the e-boxp.132Degaussing the Monitorp.1339 Troubleshootingp.134PC Won’t Bootp.134PC Boots, but Monitor is Blankp.135PC Boots, But Peripherals Don’t Workp.135Printer Problemsp.136HP940 Startupp.136Printer Installation Problemsp.136Electromagnetic Interferencep.13710 Service Proceduresp.138Overviewp.138Repair Philosophyp.139ESD Safeguardsp.139General Precautionsp.139Service Accessp.140Servicing Components Above the Lower Enclosurep.140Servicing Components in the Lower Enclosurep.140Rear Service Doorsp.141Front Service Panelp.141Side Panelsp.141Record Keepingp.141PC Servicep.142Software Service Proceduresp.142System Softwarep.142Reloading or Upgrading System Softwarep.142Backing Up and Restoring Presetsp.143Backing Up Customer Presetsp.143Restoring Customer Presetsp.146PC Componentsp.147Opening the PC Side Coverp.148Removing the PC from the Systemp.150Eboxp.151Replacing Ebox Circuit Boardsp.151Replacing the Motherboardp.152Replacing the Distribution Boardp.153Power Supplyp.154Removing the Power Supply and the Eboxp.154Separating the Ebox and the Power Supplyp.155Transducer Connector Modulesp.156Replacing Transducer Connector Modulesp.156Fanp.156Replacing the Fanp.156Monitorp.157Replacing the Monitorp.157Monitor Twivelp.158Disassembling the Twivelp.158System Control Panelp.158Replacing the System Control Panelp.159Monitor Support and Front Panelp.160Replacing the Monitor Support and its Front Panelp.160Cablesp.160System Control Panel and Monitor Cablesp.161Replacing the Upper Assembly Cable Bundlep.161Cartp.163Replacing Cart Wheels (Caster Assemblies)p.163Replacing the Top Panel of the Lower Enclosurep.164Lift Columnp.165Preparationp.165Preparing the System for Servicing the Lift Columnp.165Gas Springp.166Replacing the Gas Springp.166Mechlok and Cablep.167Replacing the Mechlok and Cablep.167Slidesp.169Replacing the Slidesp.16911 Cablingp.17112 Configurationp.174Disk Drive Settingsp.174Hard Drive Jumpersp.174CD Drive Jumpersp.175MO Drive Jumpersp.175VCR Switch Settingsp.175Sony B/W Printer Dip Switch Settingsp.17613 Partsp.177Overviewp.177Ordering Partsp.177New Partsp.178Exchange Partsp.178System Serial Number and Revision Levelp.178Major Assembliesp.178Transducer Connector Modulesp.179Internal PCp.180e-box and Power Supplyp.183Lower Assemblyp.185Lift Column Assemblyp.190Upper Assemblyp.193Top Shelfp.196VCR Option Partsp.198Softwarep.198Physio Option Partsp.199Exchange Part Numbersp.200Optional Peripheralsp.201Supplies and Accessoriesp.20214 Transducersp.20515 Glossaryp.206
Page 214 of the Philips Patient Monitor (M2540) user manual — Table of contents
Page 214 · click to enlarge
Page 214

Table of contents

1 About This Manualp.1About This Manualp.1Audiencep.2Formatp.3Conventionsp.3Service Manual Questions or Commentsp.3Customer Assistancep.41 General Informationp.6Product Overviewp.6System Descriptionp.7Cartp.7Cart Wheelsp.7PCp.7System Monitorp.8e-boxp.8System Control Panelp.8Physio Modulep.9Power Supplyp.9Optional Peripherals (VCR, Printers, and Image Devices)p.9Preset Functionsp.10Keyboard Equivalenciesp.11Safety Specificationsp.12Safety Limitsp.12Regulatory Compliancep.122 Specificationsp.13Physical Dimensionsp.14Electrical Specificationsp.19Monitorp.19Transducer Specificationsp.20Connection and Communication Specificationsp.21System I/O Portsp.21PC Portsp.21I/O Panel Portsp.23Physio Port Specificationsp.23Audio/Video Specificationsp.23Audiop.24External Videop.24Video Output Specificationsp.253 Safetyp.26Transmit Power (Acoustical)p.27Acoustic Exposurep.27AIUM/NEMA Output Display Standardp.28Automatic Index Selectionp.29When MI is Displayed With the Normal Settingp.29When TIS is Displayed With Normal Settingp.29When TIC is Displayed With Normal Settingp.30Explosive Hazardsp.30Electrical Warningsp.31Peripheral Connectionsp.31Glutaraldehyde Exposurep.32Moving the Systemp.32Before Moving the Systemp.32When Moving the Systemp.33Moving on Ramps or Uneven Surfacesp.33System Tiltingp.33Transporting the System in a Vehiclep.34Electromagnetic Compatibilityp.35Avoiding Electromagnetic Interferencep.35Restrictions for Usep.36Immunity Level Test Resultsp.36Electrostatic Dischargep.36Electrosurgical Unitsp.374 Theory of Operationp.38Overviewp.38Internal PCp.39Standard PC Componentsp.39PC Portsp.39VCR-Option Components in the PCp.40Optional Drive-Bay Componentsp.40TR Boardsp.40BPAP Boardp.42Omni TEE Circuitryp.42Demodulator Boardp.43APIO Boardp.43Physio Module Optionp.43Primary Functionsp.44Connectorsp.44System Motherboardp.44Distribution Board and Connector Modulesp.44Transducer Connector Module Typesp.45I/O Panelp.45System Control Panelp.45Power Supplyp.46Functional Block Diagramsp.47System Functional Block Diagramp.47e-box and Power Supply Functional Block Diagramp.49Token Ring Cablep.50Physio Functional Block Diagramp.51Transducer Safety Testing: Test Setup and Theoryp.52Theory of the Transducer Leakage Current Testp.545 Installationp.59Overviewp.59Assemble the Systemp.59Power Cord and Monitor Installationp.59Installing the Power Cord and Monitorp.59Peripherals Installationp.59Installing a VCRp.60Installing a Sony Color Printerp.60Installing a Sony Black-and-White Printerp.61System Startupp.62Powering On the Systemp.62Monitor Signal Voltagep.62Setting the Monitor Signal Input Voltagep.62Printer Preparationp.64Preparing the Sony Black-and-White Printerp.64Preparing the Sony Color Printerp.64System Configurationp.64Localizationp.65Setting the Localization Parametersp.65Options Installationp.71Installing Optionsp.71DICOM Network Configurationp.73Changing the System’s Network Parametersp.73Using the Field Service Options Window for Advanced Network Settingsp.74Adding Server Informationp.77Assigning Servers to DICOM Rolesp.78Configuring Autodeletep.79Configuring Loggingp.81System Hardware Configurationp.83Assigning Record Buttonsp.83Assigning Option Buttonsp.84VCR configurationp.85Configuring the System for PAL Formatp.85Setting the Monitor Color Temperaturep.87Configuring for printersp.87Customer Trainingp.886 Performance Testsp.89Overviewp.89Electrical Safety Testsp.89Chassis to Ground Resistance Testp.90Chassis to Ground Resistance Test Procedurep.91Ground Wire Leakage Current Testp.92Ground Wire Leakage Current Test Procedurep.93ECG Lead Leakage Current Test (Source)p.94ECG Lead Leakage Current Test Procedurep.95ECG Lead Isolation Leakage Current Testp.96ECG Lead Isolation Leakage Current Test Procedurep.97Transducer Testsp.98Transducer Leakage Current Test (Source)p.99Transducer Leakage Current Test Procedurep.100Transducer Isolation Leakage Current Testp.102Transducer Isolation Leakage Current Test Procedurep.103PC Diagnosticsp.105IBM Drive Fitness Testp.105Creating a DFT Bootable Floppy Diskp.105Testing PC Functions Using the Drive Fitness Testp.105Checkit Prop.106Testing PC Functions Using Checkit Prop.106Resident Self Testsp.107Availabilityp.107Logging Onp.107User Interfacep.108Information Typesp.109System Informationp.109Hardware Informationp.110Testsp.111Available Testsp.111Testing Optionsp.112Test Resultsp.113Service Logp.113Launching the RST Applicationp.113Closing the RST Applicationp.114Running a Basic Testp.114Running an Extended Testp.115Test Modulesp.117Running an Interactive Testp.118Viewing Error Logsp.119Ultrasound Application Logp.119Clearing the Ultrasound Application Logp.120RST Test Logp.121RST Log Codesp.122Clearing the RST Test Logp.122Test Result Detailsp.123Viewing Test Detailsp.123System Logp.124Clearing the System Logp.124Using the Service Logp.125Making an Entry in the Service Logp.125Modifying an Entry in the Service Logp.1267 Adjustmentsp.127Monitor Adjustmentp.127OSD Main Menup.127Opening the OSD Main Menup.127Navigating the OSD Menusp.128Adjusting VGA Vertical Sizep.128Adjust the VGA Vertical Sizep.128Changing the Boot Sequencep.130Change the Boot Sequencep.1308 Preventive Maintenancep.131Cleaning the Air Filterp.131Cleaning the Trackballp.132Cleaning the e-boxp.132Degaussing the Monitorp.1339 Troubleshootingp.134PC Won’t Bootp.134PC Boots, but Monitor is Blankp.135PC Boots, But Peripherals Don’t Workp.135Printer Problemsp.136HP940 Startupp.136Printer Installation Problemsp.136Electromagnetic Interferencep.13710 Service Proceduresp.138Overviewp.138Repair Philosophyp.139ESD Safeguardsp.139General Precautionsp.139Service Accessp.140Servicing Components Above the Lower Enclosurep.140Servicing Components in the Lower Enclosurep.140Rear Service Doorsp.141Front Service Panelp.141Side Panelsp.141Record Keepingp.141PC Servicep.142Software Service Proceduresp.142System Softwarep.142Reloading or Upgrading System Softwarep.142Backing Up and Restoring Presetsp.143Backing Up Customer Presetsp.143Restoring Customer Presetsp.146PC Componentsp.147Opening the PC Side Coverp.148Removing the PC from the Systemp.150Eboxp.151Replacing Ebox Circuit Boardsp.151Replacing the Motherboardp.152Replacing the Distribution Boardp.153Power Supplyp.154Removing the Power Supply and the Eboxp.154Separating the Ebox and the Power Supplyp.155Transducer Connector Modulesp.156Replacing Transducer Connector Modulesp.156Fanp.156Replacing the Fanp.156Monitorp.157Replacing the Monitorp.157Monitor Twivelp.158Disassembling the Twivelp.158System Control Panelp.158Replacing the System Control Panelp.159Monitor Support and Front Panelp.160Replacing the Monitor Support and its Front Panelp.160Cablesp.160System Control Panel and Monitor Cablesp.161Replacing the Upper Assembly Cable Bundlep.161Cartp.163Replacing Cart Wheels (Caster Assemblies)p.163Replacing the Top Panel of the Lower Enclosurep.164Lift Columnp.165Preparationp.165Preparing the System for Servicing the Lift Columnp.165Gas Springp.166Replacing the Gas Springp.166Mechlok and Cablep.167Replacing the Mechlok and Cablep.167Slidesp.169Replacing the Slidesp.16911 Cablingp.17112 Configurationp.174Disk Drive Settingsp.174Hard Drive Jumpersp.174CD Drive Jumpersp.175MO Drive Jumpersp.175VCR Switch Settingsp.175Sony B/W Printer Dip Switch Settingsp.17613 Partsp.177Overviewp.177Ordering Partsp.177New Partsp.178Exchange Partsp.178System Serial Number and Revision Levelp.178Major Assembliesp.178Transducer Connector Modulesp.179Internal PCp.180e-box and Power Supplyp.183Lower Assemblyp.185Lift Column Assemblyp.190Upper Assemblyp.193Top Shelfp.196VCR Option Partsp.198Softwarep.198Physio Option Partsp.199Exchange Part Numbersp.200Optional Peripheralsp.201Supplies and Accessoriesp.20214 Transducersp.20515 Glossaryp.206
Page 215 of the Philips Patient Monitor (M2540) user manual — Table of contents
Page 215 · click to enlarge
Page 215

Table of contents

1 About This Manualp.1About This Manualp.1Audiencep.2Formatp.3Conventionsp.3Service Manual Questions or Commentsp.3Customer Assistancep.41 General Informationp.6Product Overviewp.6System Descriptionp.7Cartp.7Cart Wheelsp.7PCp.7System Monitorp.8e-boxp.8System Control Panelp.8Physio Modulep.9Power Supplyp.9Optional Peripherals (VCR, Printers, and Image Devices)p.9Preset Functionsp.10Keyboard Equivalenciesp.11Safety Specificationsp.12Safety Limitsp.12Regulatory Compliancep.122 Specificationsp.13Physical Dimensionsp.14Electrical Specificationsp.19Monitorp.19Transducer Specificationsp.20Connection and Communication Specificationsp.21System I/O Portsp.21PC Portsp.21I/O Panel Portsp.23Physio Port Specificationsp.23Audio/Video Specificationsp.23Audiop.24External Videop.24Video Output Specificationsp.253 Safetyp.26Transmit Power (Acoustical)p.27Acoustic Exposurep.27AIUM/NEMA Output Display Standardp.28Automatic Index Selectionp.29When MI is Displayed With the Normal Settingp.29When TIS is Displayed With Normal Settingp.29When TIC is Displayed With Normal Settingp.30Explosive Hazardsp.30Electrical Warningsp.31Peripheral Connectionsp.31Glutaraldehyde Exposurep.32Moving the Systemp.32Before Moving the Systemp.32When Moving the Systemp.33Moving on Ramps or Uneven Surfacesp.33System Tiltingp.33Transporting the System in a Vehiclep.34Electromagnetic Compatibilityp.35Avoiding Electromagnetic Interferencep.35Restrictions for Usep.36Immunity Level Test Resultsp.36Electrostatic Dischargep.36Electrosurgical Unitsp.374 Theory of Operationp.38Overviewp.38Internal PCp.39Standard PC Componentsp.39PC Portsp.39VCR-Option Components in the PCp.40Optional Drive-Bay Componentsp.40TR Boardsp.40BPAP Boardp.42Omni TEE Circuitryp.42Demodulator Boardp.43APIO Boardp.43Physio Module Optionp.43Primary Functionsp.44Connectorsp.44System Motherboardp.44Distribution Board and Connector Modulesp.44Transducer Connector Module Typesp.45I/O Panelp.45System Control Panelp.45Power Supplyp.46Functional Block Diagramsp.47System Functional Block Diagramp.47e-box and Power Supply Functional Block Diagramp.49Token Ring Cablep.50Physio Functional Block Diagramp.51Transducer Safety Testing: Test Setup and Theoryp.52Theory of the Transducer Leakage Current Testp.545 Installationp.59Overviewp.59Assemble the Systemp.59Power Cord and Monitor Installationp.59Installing the Power Cord and Monitorp.59Peripherals Installationp.59Installing a VCRp.60Installing a Sony Color Printerp.60Installing a Sony Black-and-White Printerp.61System Startupp.62Powering On the Systemp.62Monitor Signal Voltagep.62Setting the Monitor Signal Input Voltagep.62Printer Preparationp.64Preparing the Sony Black-and-White Printerp.64Preparing the Sony Color Printerp.64System Configurationp.64Localizationp.65Setting the Localization Parametersp.65Options Installationp.71Installing Optionsp.71DICOM Network Configurationp.73Changing the System’s Network Parametersp.73Using the Field Service Options Window for Advanced Network Settingsp.74Adding Server Informationp.77Assigning Servers to DICOM Rolesp.78Configuring Autodeletep.79Configuring Loggingp.81System Hardware Configurationp.83Assigning Record Buttonsp.83Assigning Option Buttonsp.84VCR configurationp.85Configuring the System for PAL Formatp.85Setting the Monitor Color Temperaturep.87Configuring for printersp.87Customer Trainingp.886 Performance Testsp.89Overviewp.89Electrical Safety Testsp.89Chassis to Ground Resistance Testp.90Chassis to Ground Resistance Test Procedurep.91Ground Wire Leakage Current Testp.92Ground Wire Leakage Current Test Procedurep.93ECG Lead Leakage Current Test (Source)p.94ECG Lead Leakage Current Test Procedurep.95ECG Lead Isolation Leakage Current Testp.96ECG Lead Isolation Leakage Current Test Procedurep.97Transducer Testsp.98Transducer Leakage Current Test (Source)p.99Transducer Leakage Current Test Procedurep.100Transducer Isolation Leakage Current Testp.102Transducer Isolation Leakage Current Test Procedurep.103PC Diagnosticsp.105IBM Drive Fitness Testp.105Creating a DFT Bootable Floppy Diskp.105Testing PC Functions Using the Drive Fitness Testp.105Checkit Prop.106Testing PC Functions Using Checkit Prop.106Resident Self Testsp.107Availabilityp.107Logging Onp.107User Interfacep.108Information Typesp.109System Informationp.109Hardware Informationp.110Testsp.111Available Testsp.111Testing Optionsp.112Test Resultsp.113Service Logp.113Launching the RST Applicationp.113Closing the RST Applicationp.114Running a Basic Testp.114Running an Extended Testp.115Test Modulesp.117Running an Interactive Testp.118Viewing Error Logsp.119Ultrasound Application Logp.119Clearing the Ultrasound Application Logp.120RST Test Logp.121RST Log Codesp.122Clearing the RST Test Logp.122Test Result Detailsp.123Viewing Test Detailsp.123System Logp.124Clearing the System Logp.124Using the Service Logp.125Making an Entry in the Service Logp.125Modifying an Entry in the Service Logp.1267 Adjustmentsp.127Monitor Adjustmentp.127OSD Main Menup.127Opening the OSD Main Menup.127Navigating the OSD Menusp.128Adjusting VGA Vertical Sizep.128Adjust the VGA Vertical Sizep.128Changing the Boot Sequencep.130Change the Boot Sequencep.1308 Preventive Maintenancep.131Cleaning the Air Filterp.131Cleaning the Trackballp.132Cleaning the e-boxp.132Degaussing the Monitorp.1339 Troubleshootingp.134PC Won’t Bootp.134PC Boots, but Monitor is Blankp.135PC Boots, But Peripherals Don’t Workp.135Printer Problemsp.136HP940 Startupp.136Printer Installation Problemsp.136Electromagnetic Interferencep.13710 Service Proceduresp.138Overviewp.138Repair Philosophyp.139ESD Safeguardsp.139General Precautionsp.139Service Accessp.140Servicing Components Above the Lower Enclosurep.140Servicing Components in the Lower Enclosurep.140Rear Service Doorsp.141Front Service Panelp.141Side Panelsp.141Record Keepingp.141PC Servicep.142Software Service Proceduresp.142System Softwarep.142Reloading or Upgrading System Softwarep.142Backing Up and Restoring Presetsp.143Backing Up Customer Presetsp.143Restoring Customer Presetsp.146PC Componentsp.147Opening the PC Side Coverp.148Removing the PC from the Systemp.150Eboxp.151Replacing Ebox Circuit Boardsp.151Replacing the Motherboardp.152Replacing the Distribution Boardp.153Power Supplyp.154Removing the Power Supply and the Eboxp.154Separating the Ebox and the Power Supplyp.155Transducer Connector Modulesp.156Replacing Transducer Connector Modulesp.156Fanp.156Replacing the Fanp.156Monitorp.157Replacing the Monitorp.157Monitor Twivelp.158Disassembling the Twivelp.158System Control Panelp.158Replacing the System Control Panelp.159Monitor Support and Front Panelp.160Replacing the Monitor Support and its Front Panelp.160Cablesp.160System Control Panel and Monitor Cablesp.161Replacing the Upper Assembly Cable Bundlep.161Cartp.163Replacing Cart Wheels (Caster Assemblies)p.163Replacing the Top Panel of the Lower Enclosurep.164Lift Columnp.165Preparationp.165Preparing the System for Servicing the Lift Columnp.165Gas Springp.166Replacing the Gas Springp.166Mechlok and Cablep.167Replacing the Mechlok and Cablep.167Slidesp.169Replacing the Slidesp.16911 Cablingp.17112 Configurationp.174Disk Drive Settingsp.174Hard Drive Jumpersp.174CD Drive Jumpersp.175MO Drive Jumpersp.175VCR Switch Settingsp.175Sony B/W Printer Dip Switch Settingsp.17613 Partsp.177Overviewp.177Ordering Partsp.177New Partsp.178Exchange Partsp.178System Serial Number and Revision Levelp.178Major Assembliesp.178Transducer Connector Modulesp.179Internal PCp.180e-box and Power Supplyp.183Lower Assemblyp.185Lift Column Assemblyp.190Upper Assemblyp.193Top Shelfp.196VCR Option Partsp.198Softwarep.198Physio Option Partsp.199Exchange Part Numbersp.200Optional Peripheralsp.201Supplies and Accessoriesp.20214 Transducersp.20515 Glossaryp.206
Page 216 of the Philips Patient Monitor (M2540) user manual — Table of contents
Page 216 · click to enlarge
Page 216

Table of contents

1 About This Manualp.1About This Manualp.1Audiencep.2Formatp.3Conventionsp.3Service Manual Questions or Commentsp.3Customer Assistancep.41 General Informationp.6Product Overviewp.6System Descriptionp.7Cartp.7Cart Wheelsp.7PCp.7System Monitorp.8e-boxp.8System Control Panelp.8Physio Modulep.9Power Supplyp.9Optional Peripherals (VCR, Printers, and Image Devices)p.9Preset Functionsp.10Keyboard Equivalenciesp.11Safety Specificationsp.12Safety Limitsp.12Regulatory Compliancep.122 Specificationsp.13Physical Dimensionsp.14Electrical Specificationsp.19Monitorp.19Transducer Specificationsp.20Connection and Communication Specificationsp.21System I/O Portsp.21PC Portsp.21I/O Panel Portsp.23Physio Port Specificationsp.23Audio/Video Specificationsp.23Audiop.24External Videop.24Video Output Specificationsp.253 Safetyp.26Transmit Power (Acoustical)p.27Acoustic Exposurep.27AIUM/NEMA Output Display Standardp.28Automatic Index Selectionp.29When MI is Displayed With the Normal Settingp.29When TIS is Displayed With Normal Settingp.29When TIC is Displayed With Normal Settingp.30Explosive Hazardsp.30Electrical Warningsp.31Peripheral Connectionsp.31Glutaraldehyde Exposurep.32Moving the Systemp.32Before Moving the Systemp.32When Moving the Systemp.33Moving on Ramps or Uneven Surfacesp.33System Tiltingp.33Transporting the System in a Vehiclep.34Electromagnetic Compatibilityp.35Avoiding Electromagnetic Interferencep.35Restrictions for Usep.36Immunity Level Test Resultsp.36Electrostatic Dischargep.36Electrosurgical Unitsp.374 Theory of Operationp.38Overviewp.38Internal PCp.39Standard PC Componentsp.39PC Portsp.39VCR-Option Components in the PCp.40Optional Drive-Bay Componentsp.40TR Boardsp.40BPAP Boardp.42Omni TEE Circuitryp.42Demodulator Boardp.43APIO Boardp.43Physio Module Optionp.43Primary Functionsp.44Connectorsp.44System Motherboardp.44Distribution Board and Connector Modulesp.44Transducer Connector Module Typesp.45I/O Panelp.45System Control Panelp.45Power Supplyp.46Functional Block Diagramsp.47System Functional Block Diagramp.47e-box and Power Supply Functional Block Diagramp.49Token Ring Cablep.50Physio Functional Block Diagramp.51Transducer Safety Testing: Test Setup and Theoryp.52Theory of the Transducer Leakage Current Testp.545 Installationp.59Overviewp.59Assemble the Systemp.59Power Cord and Monitor Installationp.59Installing the Power Cord and Monitorp.59Peripherals Installationp.59Installing a VCRp.60Installing a Sony Color Printerp.60Installing a Sony Black-and-White Printerp.61System Startupp.62Powering On the Systemp.62Monitor Signal Voltagep.62Setting the Monitor Signal Input Voltagep.62Printer Preparationp.64Preparing the Sony Black-and-White Printerp.64Preparing the Sony Color Printerp.64System Configurationp.64Localizationp.65Setting the Localization Parametersp.65Options Installationp.71Installing Optionsp.71DICOM Network Configurationp.73Changing the System’s Network Parametersp.73Using the Field Service Options Window for Advanced Network Settingsp.74Adding Server Informationp.77Assigning Servers to DICOM Rolesp.78Configuring Autodeletep.79Configuring Loggingp.81System Hardware Configurationp.83Assigning Record Buttonsp.83Assigning Option Buttonsp.84VCR configurationp.85Configuring the System for PAL Formatp.85Setting the Monitor Color Temperaturep.87Configuring for printersp.87Customer Trainingp.886 Performance Testsp.89Overviewp.89Electrical Safety Testsp.89Chassis to Ground Resistance Testp.90Chassis to Ground Resistance Test Procedurep.91Ground Wire Leakage Current Testp.92Ground Wire Leakage Current Test Procedurep.93ECG Lead Leakage Current Test (Source)p.94ECG Lead Leakage Current Test Procedurep.95ECG Lead Isolation Leakage Current Testp.96ECG Lead Isolation Leakage Current Test Procedurep.97Transducer Testsp.98Transducer Leakage Current Test (Source)p.99Transducer Leakage Current Test Procedurep.100Transducer Isolation Leakage Current Testp.102Transducer Isolation Leakage Current Test Procedurep.103PC Diagnosticsp.105IBM Drive Fitness Testp.105Creating a DFT Bootable Floppy Diskp.105Testing PC Functions Using the Drive Fitness Testp.105Checkit Prop.106Testing PC Functions Using Checkit Prop.106Resident Self Testsp.107Availabilityp.107Logging Onp.107User Interfacep.108Information Typesp.109System Informationp.109Hardware Informationp.110Testsp.111Available Testsp.111Testing Optionsp.112Test Resultsp.113Service Logp.113Launching the RST Applicationp.113Closing the RST Applicationp.114Running a Basic Testp.114Running an Extended Testp.115Test Modulesp.117Running an Interactive Testp.118Viewing Error Logsp.119Ultrasound Application Logp.119Clearing the Ultrasound Application Logp.120RST Test Logp.121RST Log Codesp.122Clearing the RST Test Logp.122Test Result Detailsp.123Viewing Test Detailsp.123System Logp.124Clearing the System Logp.124Using the Service Logp.125Making an Entry in the Service Logp.125Modifying an Entry in the Service Logp.1267 Adjustmentsp.127Monitor Adjustmentp.127OSD Main Menup.127Opening the OSD Main Menup.127Navigating the OSD Menusp.128Adjusting VGA Vertical Sizep.128Adjust the VGA Vertical Sizep.128Changing the Boot Sequencep.130Change the Boot Sequencep.1308 Preventive Maintenancep.131Cleaning the Air Filterp.131Cleaning the Trackballp.132Cleaning the e-boxp.132Degaussing the Monitorp.1339 Troubleshootingp.134PC Won’t Bootp.134PC Boots, but Monitor is Blankp.135PC Boots, But Peripherals Don’t Workp.135Printer Problemsp.136HP940 Startupp.136Printer Installation Problemsp.136Electromagnetic Interferencep.13710 Service Proceduresp.138Overviewp.138Repair Philosophyp.139ESD Safeguardsp.139General Precautionsp.139Service Accessp.140Servicing Components Above the Lower Enclosurep.140Servicing Components in the Lower Enclosurep.140Rear Service Doorsp.141Front Service Panelp.141Side Panelsp.141Record Keepingp.141PC Servicep.142Software Service Proceduresp.142System Softwarep.142Reloading or Upgrading System Softwarep.142Backing Up and Restoring Presetsp.143Backing Up Customer Presetsp.143Restoring Customer Presetsp.146PC Componentsp.147Opening the PC Side Coverp.148Removing the PC from the Systemp.150Eboxp.151Replacing Ebox Circuit Boardsp.151Replacing the Motherboardp.152Replacing the Distribution Boardp.153Power Supplyp.154Removing the Power Supply and the Eboxp.154Separating the Ebox and the Power Supplyp.155Transducer Connector Modulesp.156Replacing Transducer Connector Modulesp.156Fanp.156Replacing the Fanp.156Monitorp.157Replacing the Monitorp.157Monitor Twivelp.158Disassembling the Twivelp.158System Control Panelp.158Replacing the System Control Panelp.159Monitor Support and Front Panelp.160Replacing the Monitor Support and its Front Panelp.160Cablesp.160System Control Panel and Monitor Cablesp.161Replacing the Upper Assembly Cable Bundlep.161Cartp.163Replacing Cart Wheels (Caster Assemblies)p.163Replacing the Top Panel of the Lower Enclosurep.164Lift Columnp.165Preparationp.165Preparing the System for Servicing the Lift Columnp.165Gas Springp.166Replacing the Gas Springp.166Mechlok and Cablep.167Replacing the Mechlok and Cablep.167Slidesp.169Replacing the Slidesp.16911 Cablingp.17112 Configurationp.174Disk Drive Settingsp.174Hard Drive Jumpersp.174CD Drive Jumpersp.175MO Drive Jumpersp.175VCR Switch Settingsp.175Sony B/W Printer Dip Switch Settingsp.17613 Partsp.177Overviewp.177Ordering Partsp.177New Partsp.178Exchange Partsp.178System Serial Number and Revision Levelp.178Major Assembliesp.178Transducer Connector Modulesp.179Internal PCp.180e-box and Power Supplyp.183Lower Assemblyp.185Lift Column Assemblyp.190Upper Assemblyp.193Top Shelfp.196VCR Option Partsp.198Softwarep.198Physio Option Partsp.199Exchange Part Numbersp.200Optional Peripheralsp.201Supplies and Accessoriesp.20214 Transducersp.20515 Glossaryp.206
Page 217 of the Philips Patient Monitor (M2540) user manual — Table of contents
Page 217 · click to enlarge
Page 217

Table of contents

1 About This Manualp.1About This Manualp.1Audiencep.2Formatp.3Conventionsp.3Service Manual Questions or Commentsp.3Customer Assistancep.41 General Informationp.6Product Overviewp.6System Descriptionp.7Cartp.7Cart Wheelsp.7PCp.7System Monitorp.8e-boxp.8System Control Panelp.8Physio Modulep.9Power Supplyp.9Optional Peripherals (VCR, Printers, and Image Devices)p.9Preset Functionsp.10Keyboard Equivalenciesp.11Safety Specificationsp.12Safety Limitsp.12Regulatory Compliancep.122 Specificationsp.13Physical Dimensionsp.14Electrical Specificationsp.19Monitorp.19Transducer Specificationsp.20Connection and Communication Specificationsp.21System I/O Portsp.21PC Portsp.21I/O Panel Portsp.23Physio Port Specificationsp.23Audio/Video Specificationsp.23Audiop.24External Videop.24Video Output Specificationsp.253 Safetyp.26Transmit Power (Acoustical)p.27Acoustic Exposurep.27AIUM/NEMA Output Display Standardp.28Automatic Index Selectionp.29When MI is Displayed With the Normal Settingp.29When TIS is Displayed With Normal Settingp.29When TIC is Displayed With Normal Settingp.30Explosive Hazardsp.30Electrical Warningsp.31Peripheral Connectionsp.31Glutaraldehyde Exposurep.32Moving the Systemp.32Before Moving the Systemp.32When Moving the Systemp.33Moving on Ramps or Uneven Surfacesp.33System Tiltingp.33Transporting the System in a Vehiclep.34Electromagnetic Compatibilityp.35Avoiding Electromagnetic Interferencep.35Restrictions for Usep.36Immunity Level Test Resultsp.36Electrostatic Dischargep.36Electrosurgical Unitsp.374 Theory of Operationp.38Overviewp.38Internal PCp.39Standard PC Componentsp.39PC Portsp.39VCR-Option Components in the PCp.40Optional Drive-Bay Componentsp.40TR Boardsp.40BPAP Boardp.42Omni TEE Circuitryp.42Demodulator Boardp.43APIO Boardp.43Physio Module Optionp.43Primary Functionsp.44Connectorsp.44System Motherboardp.44Distribution Board and Connector Modulesp.44Transducer Connector Module Typesp.45I/O Panelp.45System Control Panelp.45Power Supplyp.46Functional Block Diagramsp.47System Functional Block Diagramp.47e-box and Power Supply Functional Block Diagramp.49Token Ring Cablep.50Physio Functional Block Diagramp.51Transducer Safety Testing: Test Setup and Theoryp.52Theory of the Transducer Leakage Current Testp.545 Installationp.59Overviewp.59Assemble the Systemp.59Power Cord and Monitor Installationp.59Installing the Power Cord and Monitorp.59Peripherals Installationp.59Installing a VCRp.60Installing a Sony Color Printerp.60Installing a Sony Black-and-White Printerp.61System Startupp.62Powering On the Systemp.62Monitor Signal Voltagep.62Setting the Monitor Signal Input Voltagep.62Printer Preparationp.64Preparing the Sony Black-and-White Printerp.64Preparing the Sony Color Printerp.64System Configurationp.64Localizationp.65Setting the Localization Parametersp.65Options Installationp.71Installing Optionsp.71DICOM Network Configurationp.73Changing the System’s Network Parametersp.73Using the Field Service Options Window for Advanced Network Settingsp.74Adding Server Informationp.77Assigning Servers to DICOM Rolesp.78Configuring Autodeletep.79Configuring Loggingp.81System Hardware Configurationp.83Assigning Record Buttonsp.83Assigning Option Buttonsp.84VCR configurationp.85Configuring the System for PAL Formatp.85Setting the Monitor Color Temperaturep.87Configuring for printersp.87Customer Trainingp.886 Performance Testsp.89Overviewp.89Electrical Safety Testsp.89Chassis to Ground Resistance Testp.90Chassis to Ground Resistance Test Procedurep.91Ground Wire Leakage Current Testp.92Ground Wire Leakage Current Test Procedurep.93ECG Lead Leakage Current Test (Source)p.94ECG Lead Leakage Current Test Procedurep.95ECG Lead Isolation Leakage Current Testp.96ECG Lead Isolation Leakage Current Test Procedurep.97Transducer Testsp.98Transducer Leakage Current Test (Source)p.99Transducer Leakage Current Test Procedurep.100Transducer Isolation Leakage Current Testp.102Transducer Isolation Leakage Current Test Procedurep.103PC Diagnosticsp.105IBM Drive Fitness Testp.105Creating a DFT Bootable Floppy Diskp.105Testing PC Functions Using the Drive Fitness Testp.105Checkit Prop.106Testing PC Functions Using Checkit Prop.106Resident Self Testsp.107Availabilityp.107Logging Onp.107User Interfacep.108Information Typesp.109System Informationp.109Hardware Informationp.110Testsp.111Available Testsp.111Testing Optionsp.112Test Resultsp.113Service Logp.113Launching the RST Applicationp.113Closing the RST Applicationp.114Running a Basic Testp.114Running an Extended Testp.115Test Modulesp.117Running an Interactive Testp.118Viewing Error Logsp.119Ultrasound Application Logp.119Clearing the Ultrasound Application Logp.120RST Test Logp.121RST Log Codesp.122Clearing the RST Test Logp.122Test Result Detailsp.123Viewing Test Detailsp.123System Logp.124Clearing the System Logp.124Using the Service Logp.125Making an Entry in the Service Logp.125Modifying an Entry in the Service Logp.1267 Adjustmentsp.127Monitor Adjustmentp.127OSD Main Menup.127Opening the OSD Main Menup.127Navigating the OSD Menusp.128Adjusting VGA Vertical Sizep.128Adjust the VGA Vertical Sizep.128Changing the Boot Sequencep.130Change the Boot Sequencep.1308 Preventive Maintenancep.131Cleaning the Air Filterp.131Cleaning the Trackballp.132Cleaning the e-boxp.132Degaussing the Monitorp.1339 Troubleshootingp.134PC Won’t Bootp.134PC Boots, but Monitor is Blankp.135PC Boots, But Peripherals Don’t Workp.135Printer Problemsp.136HP940 Startupp.136Printer Installation Problemsp.136Electromagnetic Interferencep.13710 Service Proceduresp.138Overviewp.138Repair Philosophyp.139ESD Safeguardsp.139General Precautionsp.139Service Accessp.140Servicing Components Above the Lower Enclosurep.140Servicing Components in the Lower Enclosurep.140Rear Service Doorsp.141Front Service Panelp.141Side Panelsp.141Record Keepingp.141PC Servicep.142Software Service Proceduresp.142System Softwarep.142Reloading or Upgrading System Softwarep.142Backing Up and Restoring Presetsp.143Backing Up Customer Presetsp.143Restoring Customer Presetsp.146PC Componentsp.147Opening the PC Side Coverp.148Removing the PC from the Systemp.150Eboxp.151Replacing Ebox Circuit Boardsp.151Replacing the Motherboardp.152Replacing the Distribution Boardp.153Power Supplyp.154Removing the Power Supply and the Eboxp.154Separating the Ebox and the Power Supplyp.155Transducer Connector Modulesp.156Replacing Transducer Connector Modulesp.156Fanp.156Replacing the Fanp.156Monitorp.157Replacing the Monitorp.157Monitor Twivelp.158Disassembling the Twivelp.158System Control Panelp.158Replacing the System Control Panelp.159Monitor Support and Front Panelp.160Replacing the Monitor Support and its Front Panelp.160Cablesp.160System Control Panel and Monitor Cablesp.161Replacing the Upper Assembly Cable Bundlep.161Cartp.163Replacing Cart Wheels (Caster Assemblies)p.163Replacing the Top Panel of the Lower Enclosurep.164Lift Columnp.165Preparationp.165Preparing the System for Servicing the Lift Columnp.165Gas Springp.166Replacing the Gas Springp.166Mechlok and Cablep.167Replacing the Mechlok and Cablep.167Slidesp.169Replacing the Slidesp.16911 Cablingp.17112 Configurationp.174Disk Drive Settingsp.174Hard Drive Jumpersp.174CD Drive Jumpersp.175MO Drive Jumpersp.175VCR Switch Settingsp.175Sony B/W Printer Dip Switch Settingsp.17613 Partsp.177Overviewp.177Ordering Partsp.177New Partsp.178Exchange Partsp.178System Serial Number and Revision Levelp.178Major Assembliesp.178Transducer Connector Modulesp.179Internal PCp.180e-box and Power Supplyp.183Lower Assemblyp.185Lift Column Assemblyp.190Upper Assemblyp.193Top Shelfp.196VCR Option Partsp.198Softwarep.198Physio Option Partsp.199Exchange Part Numbersp.200Optional Peripheralsp.201Supplies and Accessoriesp.20214 Transducersp.20515 Glossaryp.206
Page 218 of the Philips Patient Monitor (M2540) user manual — Table of contents
Page 218 · click to enlarge
Page 218

Table of contents

1 About This Manualp.1About This Manualp.1Audiencep.2Formatp.3Conventionsp.3Service Manual Questions or Commentsp.3Customer Assistancep.41 General Informationp.6Product Overviewp.6System Descriptionp.7Cartp.7Cart Wheelsp.7PCp.7System Monitorp.8e-boxp.8System Control Panelp.8Physio Modulep.9Power Supplyp.9Optional Peripherals (VCR, Printers, and Image Devices)p.9Preset Functionsp.10Keyboard Equivalenciesp.11Safety Specificationsp.12Safety Limitsp.12Regulatory Compliancep.122 Specificationsp.13Physical Dimensionsp.14Electrical Specificationsp.19Monitorp.19Transducer Specificationsp.20Connection and Communication Specificationsp.21System I/O Portsp.21PC Portsp.21I/O Panel Portsp.23Physio Port Specificationsp.23Audio/Video Specificationsp.23Audiop.24External Videop.24Video Output Specificationsp.253 Safetyp.26Transmit Power (Acoustical)p.27Acoustic Exposurep.27AIUM/NEMA Output Display Standardp.28Automatic Index Selectionp.29When MI is Displayed With the Normal Settingp.29When TIS is Displayed With Normal Settingp.29When TIC is Displayed With Normal Settingp.30Explosive Hazardsp.30Electrical Warningsp.31Peripheral Connectionsp.31Glutaraldehyde Exposurep.32Moving the Systemp.32Before Moving the Systemp.32When Moving the Systemp.33Moving on Ramps or Uneven Surfacesp.33System Tiltingp.33Transporting the System in a Vehiclep.34Electromagnetic Compatibilityp.35Avoiding Electromagnetic Interferencep.35Restrictions for Usep.36Immunity Level Test Resultsp.36Electrostatic Dischargep.36Electrosurgical Unitsp.374 Theory of Operationp.38Overviewp.38Internal PCp.39Standard PC Componentsp.39PC Portsp.39VCR-Option Components in the PCp.40Optional Drive-Bay Componentsp.40TR Boardsp.40BPAP Boardp.42Omni TEE Circuitryp.42Demodulator Boardp.43APIO Boardp.43Physio Module Optionp.43Primary Functionsp.44Connectorsp.44System Motherboardp.44Distribution Board and Connector Modulesp.44Transducer Connector Module Typesp.45I/O Panelp.45System Control Panelp.45Power Supplyp.46Functional Block Diagramsp.47System Functional Block Diagramp.47e-box and Power Supply Functional Block Diagramp.49Token Ring Cablep.50Physio Functional Block Diagramp.51Transducer Safety Testing: Test Setup and Theoryp.52Theory of the Transducer Leakage Current Testp.545 Installationp.59Overviewp.59Assemble the Systemp.59Power Cord and Monitor Installationp.59Installing the Power Cord and Monitorp.59Peripherals Installationp.59Installing a VCRp.60Installing a Sony Color Printerp.60Installing a Sony Black-and-White Printerp.61System Startupp.62Powering On the Systemp.62Monitor Signal Voltagep.62Setting the Monitor Signal Input Voltagep.62Printer Preparationp.64Preparing the Sony Black-and-White Printerp.64Preparing the Sony Color Printerp.64System Configurationp.64Localizationp.65Setting the Localization Parametersp.65Options Installationp.71Installing Optionsp.71DICOM Network Configurationp.73Changing the System’s Network Parametersp.73Using the Field Service Options Window for Advanced Network Settingsp.74Adding Server Informationp.77Assigning Servers to DICOM Rolesp.78Configuring Autodeletep.79Configuring Loggingp.81System Hardware Configurationp.83Assigning Record Buttonsp.83Assigning Option Buttonsp.84VCR configurationp.85Configuring the System for PAL Formatp.85Setting the Monitor Color Temperaturep.87Configuring for printersp.87Customer Trainingp.886 Performance Testsp.89Overviewp.89Electrical Safety Testsp.89Chassis to Ground Resistance Testp.90Chassis to Ground Resistance Test Procedurep.91Ground Wire Leakage Current Testp.92Ground Wire Leakage Current Test Procedurep.93ECG Lead Leakage Current Test (Source)p.94ECG Lead Leakage Current Test Procedurep.95ECG Lead Isolation Leakage Current Testp.96ECG Lead Isolation Leakage Current Test Procedurep.97Transducer Testsp.98Transducer Leakage Current Test (Source)p.99Transducer Leakage Current Test Procedurep.100Transducer Isolation Leakage Current Testp.102Transducer Isolation Leakage Current Test Procedurep.103PC Diagnosticsp.105IBM Drive Fitness Testp.105Creating a DFT Bootable Floppy Diskp.105Testing PC Functions Using the Drive Fitness Testp.105Checkit Prop.106Testing PC Functions Using Checkit Prop.106Resident Self Testsp.107Availabilityp.107Logging Onp.107User Interfacep.108Information Typesp.109System Informationp.109Hardware Informationp.110Testsp.111Available Testsp.111Testing Optionsp.112Test Resultsp.113Service Logp.113Launching the RST Applicationp.113Closing the RST Applicationp.114Running a Basic Testp.114Running an Extended Testp.115Test Modulesp.117Running an Interactive Testp.118Viewing Error Logsp.119Ultrasound Application Logp.119Clearing the Ultrasound Application Logp.120RST Test Logp.121RST Log Codesp.122Clearing the RST Test Logp.122Test Result Detailsp.123Viewing Test Detailsp.123System Logp.124Clearing the System Logp.124Using the Service Logp.125Making an Entry in the Service Logp.125Modifying an Entry in the Service Logp.1267 Adjustmentsp.127Monitor Adjustmentp.127OSD Main Menup.127Opening the OSD Main Menup.127Navigating the OSD Menusp.128Adjusting VGA Vertical Sizep.128Adjust the VGA Vertical Sizep.128Changing the Boot Sequencep.130Change the Boot Sequencep.1308 Preventive Maintenancep.131Cleaning the Air Filterp.131Cleaning the Trackballp.132Cleaning the e-boxp.132Degaussing the Monitorp.1339 Troubleshootingp.134PC Won’t Bootp.134PC Boots, but Monitor is Blankp.135PC Boots, But Peripherals Don’t Workp.135Printer Problemsp.136HP940 Startupp.136Printer Installation Problemsp.136Electromagnetic Interferencep.13710 Service Proceduresp.138Overviewp.138Repair Philosophyp.139ESD Safeguardsp.139General Precautionsp.139Service Accessp.140Servicing Components Above the Lower Enclosurep.140Servicing Components in the Lower Enclosurep.140Rear Service Doorsp.141Front Service Panelp.141Side Panelsp.141Record Keepingp.141PC Servicep.142Software Service Proceduresp.142System Softwarep.142Reloading or Upgrading System Softwarep.142Backing Up and Restoring Presetsp.143Backing Up Customer Presetsp.143Restoring Customer Presetsp.146PC Componentsp.147Opening the PC Side Coverp.148Removing the PC from the Systemp.150Eboxp.151Replacing Ebox Circuit Boardsp.151Replacing the Motherboardp.152Replacing the Distribution Boardp.153Power Supplyp.154Removing the Power Supply and the Eboxp.154Separating the Ebox and the Power Supplyp.155Transducer Connector Modulesp.156Replacing Transducer Connector Modulesp.156Fanp.156Replacing the Fanp.156Monitorp.157Replacing the Monitorp.157Monitor Twivelp.158Disassembling the Twivelp.158System Control Panelp.158Replacing the System Control Panelp.159Monitor Support and Front Panelp.160Replacing the Monitor Support and its Front Panelp.160Cablesp.160System Control Panel and Monitor Cablesp.161Replacing the Upper Assembly Cable Bundlep.161Cartp.163Replacing Cart Wheels (Caster Assemblies)p.163Replacing the Top Panel of the Lower Enclosurep.164Lift Columnp.165Preparationp.165Preparing the System for Servicing the Lift Columnp.165Gas Springp.166Replacing the Gas Springp.166Mechlok and Cablep.167Replacing the Mechlok and Cablep.167Slidesp.169Replacing the Slidesp.16911 Cablingp.17112 Configurationp.174Disk Drive Settingsp.174Hard Drive Jumpersp.174CD Drive Jumpersp.175MO Drive Jumpersp.175VCR Switch Settingsp.175Sony B/W Printer Dip Switch Settingsp.17613 Partsp.177Overviewp.177Ordering Partsp.177New Partsp.178Exchange Partsp.178System Serial Number and Revision Levelp.178Major Assembliesp.178Transducer Connector Modulesp.179Internal PCp.180e-box and Power Supplyp.183Lower Assemblyp.185Lift Column Assemblyp.190Upper Assemblyp.193Top Shelfp.196VCR Option Partsp.198Softwarep.198Physio Option Partsp.199Exchange Part Numbersp.200Optional Peripheralsp.201Supplies and Accessoriesp.20214 Transducersp.20515 Glossaryp.206
Page 219 of the Philips Patient Monitor (M2540) user manual — Table of contents
Page 219 · click to enlarge
Page 219

Table of contents

1 About This Manualp.1About This Manualp.1Audiencep.2Formatp.3Conventionsp.3Service Manual Questions or Commentsp.3Customer Assistancep.41 General Informationp.6Product Overviewp.6System Descriptionp.7Cartp.7Cart Wheelsp.7PCp.7System Monitorp.8e-boxp.8System Control Panelp.8Physio Modulep.9Power Supplyp.9Optional Peripherals (VCR, Printers, and Image Devices)p.9Preset Functionsp.10Keyboard Equivalenciesp.11Safety Specificationsp.12Safety Limitsp.12Regulatory Compliancep.122 Specificationsp.13Physical Dimensionsp.14Electrical Specificationsp.19Monitorp.19Transducer Specificationsp.20Connection and Communication Specificationsp.21System I/O Portsp.21PC Portsp.21I/O Panel Portsp.23Physio Port Specificationsp.23Audio/Video Specificationsp.23Audiop.24External Videop.24Video Output Specificationsp.253 Safetyp.26Transmit Power (Acoustical)p.27Acoustic Exposurep.27AIUM/NEMA Output Display Standardp.28Automatic Index Selectionp.29When MI is Displayed With the Normal Settingp.29When TIS is Displayed With Normal Settingp.29When TIC is Displayed With Normal Settingp.30Explosive Hazardsp.30Electrical Warningsp.31Peripheral Connectionsp.31Glutaraldehyde Exposurep.32Moving the Systemp.32Before Moving the Systemp.32When Moving the Systemp.33Moving on Ramps or Uneven Surfacesp.33System Tiltingp.33Transporting the System in a Vehiclep.34Electromagnetic Compatibilityp.35Avoiding Electromagnetic Interferencep.35Restrictions for Usep.36Immunity Level Test Resultsp.36Electrostatic Dischargep.36Electrosurgical Unitsp.374 Theory of Operationp.38Overviewp.38Internal PCp.39Standard PC Componentsp.39PC Portsp.39VCR-Option Components in the PCp.40Optional Drive-Bay Componentsp.40TR Boardsp.40BPAP Boardp.42Omni TEE Circuitryp.42Demodulator Boardp.43APIO Boardp.43Physio Module Optionp.43Primary Functionsp.44Connectorsp.44System Motherboardp.44Distribution Board and Connector Modulesp.44Transducer Connector Module Typesp.45I/O Panelp.45System Control Panelp.45Power Supplyp.46Functional Block Diagramsp.47System Functional Block Diagramp.47e-box and Power Supply Functional Block Diagramp.49Token Ring Cablep.50Physio Functional Block Diagramp.51Transducer Safety Testing: Test Setup and Theoryp.52Theory of the Transducer Leakage Current Testp.545 Installationp.59Overviewp.59Assemble the Systemp.59Power Cord and Monitor Installationp.59Installing the Power Cord and Monitorp.59Peripherals Installationp.59Installing a VCRp.60Installing a Sony Color Printerp.60Installing a Sony Black-and-White Printerp.61System Startupp.62Powering On the Systemp.62Monitor Signal Voltagep.62Setting the Monitor Signal Input Voltagep.62Printer Preparationp.64Preparing the Sony Black-and-White Printerp.64Preparing the Sony Color Printerp.64System Configurationp.64Localizationp.65Setting the Localization Parametersp.65Options Installationp.71Installing Optionsp.71DICOM Network Configurationp.73Changing the System’s Network Parametersp.73Using the Field Service Options Window for Advanced Network Settingsp.74Adding Server Informationp.77Assigning Servers to DICOM Rolesp.78Configuring Autodeletep.79Configuring Loggingp.81System Hardware Configurationp.83Assigning Record Buttonsp.83Assigning Option Buttonsp.84VCR configurationp.85Configuring the System for PAL Formatp.85Setting the Monitor Color Temperaturep.87Configuring for printersp.87Customer Trainingp.886 Performance Testsp.89Overviewp.89Electrical Safety Testsp.89Chassis to Ground Resistance Testp.90Chassis to Ground Resistance Test Procedurep.91Ground Wire Leakage Current Testp.92Ground Wire Leakage Current Test Procedurep.93ECG Lead Leakage Current Test (Source)p.94ECG Lead Leakage Current Test Procedurep.95ECG Lead Isolation Leakage Current Testp.96ECG Lead Isolation Leakage Current Test Procedurep.97Transducer Testsp.98Transducer Leakage Current Test (Source)p.99Transducer Leakage Current Test Procedurep.100Transducer Isolation Leakage Current Testp.102Transducer Isolation Leakage Current Test Procedurep.103PC Diagnosticsp.105IBM Drive Fitness Testp.105Creating a DFT Bootable Floppy Diskp.105Testing PC Functions Using the Drive Fitness Testp.105Checkit Prop.106Testing PC Functions Using Checkit Prop.106Resident Self Testsp.107Availabilityp.107Logging Onp.107User Interfacep.108Information Typesp.109System Informationp.109Hardware Informationp.110Testsp.111Available Testsp.111Testing Optionsp.112Test Resultsp.113Service Logp.113Launching the RST Applicationp.113Closing the RST Applicationp.114Running a Basic Testp.114Running an Extended Testp.115Test Modulesp.117Running an Interactive Testp.118Viewing Error Logsp.119Ultrasound Application Logp.119Clearing the Ultrasound Application Logp.120RST Test Logp.121RST Log Codesp.122Clearing the RST Test Logp.122Test Result Detailsp.123Viewing Test Detailsp.123System Logp.124Clearing the System Logp.124Using the Service Logp.125Making an Entry in the Service Logp.125Modifying an Entry in the Service Logp.1267 Adjustmentsp.127Monitor Adjustmentp.127OSD Main Menup.127Opening the OSD Main Menup.127Navigating the OSD Menusp.128Adjusting VGA Vertical Sizep.128Adjust the VGA Vertical Sizep.128Changing the Boot Sequencep.130Change the Boot Sequencep.1308 Preventive Maintenancep.131Cleaning the Air Filterp.131Cleaning the Trackballp.132Cleaning the e-boxp.132Degaussing the Monitorp.1339 Troubleshootingp.134PC Won’t Bootp.134PC Boots, but Monitor is Blankp.135PC Boots, But Peripherals Don’t Workp.135Printer Problemsp.136HP940 Startupp.136Printer Installation Problemsp.136Electromagnetic Interferencep.13710 Service Proceduresp.138Overviewp.138Repair Philosophyp.139ESD Safeguardsp.139General Precautionsp.139Service Accessp.140Servicing Components Above the Lower Enclosurep.140Servicing Components in the Lower Enclosurep.140Rear Service Doorsp.141Front Service Panelp.141Side Panelsp.141Record Keepingp.141PC Servicep.142Software Service Proceduresp.142System Softwarep.142Reloading or Upgrading System Softwarep.142Backing Up and Restoring Presetsp.143Backing Up Customer Presetsp.143Restoring Customer Presetsp.146PC Componentsp.147Opening the PC Side Coverp.148Removing the PC from the Systemp.150Eboxp.151Replacing Ebox Circuit Boardsp.151Replacing the Motherboardp.152Replacing the Distribution Boardp.153Power Supplyp.154Removing the Power Supply and the Eboxp.154Separating the Ebox and the Power Supplyp.155Transducer Connector Modulesp.156Replacing Transducer Connector Modulesp.156Fanp.156Replacing the Fanp.156Monitorp.157Replacing the Monitorp.157Monitor Twivelp.158Disassembling the Twivelp.158System Control Panelp.158Replacing the System Control Panelp.159Monitor Support and Front Panelp.160Replacing the Monitor Support and its Front Panelp.160Cablesp.160System Control Panel and Monitor Cablesp.161Replacing the Upper Assembly Cable Bundlep.161Cartp.163Replacing Cart Wheels (Caster Assemblies)p.163Replacing the Top Panel of the Lower Enclosurep.164Lift Columnp.165Preparationp.165Preparing the System for Servicing the Lift Columnp.165Gas Springp.166Replacing the Gas Springp.166Mechlok and Cablep.167Replacing the Mechlok and Cablep.167Slidesp.169Replacing the Slidesp.16911 Cablingp.17112 Configurationp.174Disk Drive Settingsp.174Hard Drive Jumpersp.174CD Drive Jumpersp.175MO Drive Jumpersp.175VCR Switch Settingsp.175Sony B/W Printer Dip Switch Settingsp.17613 Partsp.177Overviewp.177Ordering Partsp.177New Partsp.178Exchange Partsp.178System Serial Number and Revision Levelp.178Major Assembliesp.178Transducer Connector Modulesp.179Internal PCp.180e-box and Power Supplyp.183Lower Assemblyp.185Lift Column Assemblyp.190Upper Assemblyp.193Top Shelfp.196VCR Option Partsp.198Softwarep.198Physio Option Partsp.199Exchange Part Numbersp.200Optional Peripheralsp.201Supplies and Accessoriesp.20214 Transducersp.20515 Glossaryp.206
Page 220 of the Philips Patient Monitor (M2540) user manual — Table of contents
Page 220 · click to enlarge
Page 220

Table of contents

1 About This Manualp.1About This Manualp.1Audiencep.2Formatp.3Conventionsp.3Service Manual Questions or Commentsp.3Customer Assistancep.41 General Informationp.6Product Overviewp.6System Descriptionp.7Cartp.7Cart Wheelsp.7PCp.7System Monitorp.8e-boxp.8System Control Panelp.8Physio Modulep.9Power Supplyp.9Optional Peripherals (VCR, Printers, and Image Devices)p.9Preset Functionsp.10Keyboard Equivalenciesp.11Safety Specificationsp.12Safety Limitsp.12Regulatory Compliancep.122 Specificationsp.13Physical Dimensionsp.14Electrical Specificationsp.19Monitorp.19Transducer Specificationsp.20Connection and Communication Specificationsp.21System I/O Portsp.21PC Portsp.21I/O Panel Portsp.23Physio Port Specificationsp.23Audio/Video Specificationsp.23Audiop.24External Videop.24Video Output Specificationsp.253 Safetyp.26Transmit Power (Acoustical)p.27Acoustic Exposurep.27AIUM/NEMA Output Display Standardp.28Automatic Index Selectionp.29When MI is Displayed With the Normal Settingp.29When TIS is Displayed With Normal Settingp.29When TIC is Displayed With Normal Settingp.30Explosive Hazardsp.30Electrical Warningsp.31Peripheral Connectionsp.31Glutaraldehyde Exposurep.32Moving the Systemp.32Before Moving the Systemp.32When Moving the Systemp.33Moving on Ramps or Uneven Surfacesp.33System Tiltingp.33Transporting the System in a Vehiclep.34Electromagnetic Compatibilityp.35Avoiding Electromagnetic Interferencep.35Restrictions for Usep.36Immunity Level Test Resultsp.36Electrostatic Dischargep.36Electrosurgical Unitsp.374 Theory of Operationp.38Overviewp.38Internal PCp.39Standard PC Componentsp.39PC Portsp.39VCR-Option Components in the PCp.40Optional Drive-Bay Componentsp.40TR Boardsp.40BPAP Boardp.42Omni TEE Circuitryp.42Demodulator Boardp.43APIO Boardp.43Physio Module Optionp.43Primary Functionsp.44Connectorsp.44System Motherboardp.44Distribution Board and Connector Modulesp.44Transducer Connector Module Typesp.45I/O Panelp.45System Control Panelp.45Power Supplyp.46Functional Block Diagramsp.47System Functional Block Diagramp.47e-box and Power Supply Functional Block Diagramp.49Token Ring Cablep.50Physio Functional Block Diagramp.51Transducer Safety Testing: Test Setup and Theoryp.52Theory of the Transducer Leakage Current Testp.545 Installationp.59Overviewp.59Assemble the Systemp.59Power Cord and Monitor Installationp.59Installing the Power Cord and Monitorp.59Peripherals Installationp.59Installing a VCRp.60Installing a Sony Color Printerp.60Installing a Sony Black-and-White Printerp.61System Startupp.62Powering On the Systemp.62Monitor Signal Voltagep.62Setting the Monitor Signal Input Voltagep.62Printer Preparationp.64Preparing the Sony Black-and-White Printerp.64Preparing the Sony Color Printerp.64System Configurationp.64Localizationp.65Setting the Localization Parametersp.65Options Installationp.71Installing Optionsp.71DICOM Network Configurationp.73Changing the System’s Network Parametersp.73Using the Field Service Options Window for Advanced Network Settingsp.74Adding Server Informationp.77Assigning Servers to DICOM Rolesp.78Configuring Autodeletep.79Configuring Loggingp.81System Hardware Configurationp.83Assigning Record Buttonsp.83Assigning Option Buttonsp.84VCR configurationp.85Configuring the System for PAL Formatp.85Setting the Monitor Color Temperaturep.87Configuring for printersp.87Customer Trainingp.886 Performance Testsp.89Overviewp.89Electrical Safety Testsp.89Chassis to Ground Resistance Testp.90Chassis to Ground Resistance Test Procedurep.91Ground Wire Leakage Current Testp.92Ground Wire Leakage Current Test Procedurep.93ECG Lead Leakage Current Test (Source)p.94ECG Lead Leakage Current Test Procedurep.95ECG Lead Isolation Leakage Current Testp.96ECG Lead Isolation Leakage Current Test Procedurep.97Transducer Testsp.98Transducer Leakage Current Test (Source)p.99Transducer Leakage Current Test Procedurep.100Transducer Isolation Leakage Current Testp.102Transducer Isolation Leakage Current Test Procedurep.103PC Diagnosticsp.105IBM Drive Fitness Testp.105Creating a DFT Bootable Floppy Diskp.105Testing PC Functions Using the Drive Fitness Testp.105Checkit Prop.106Testing PC Functions Using Checkit Prop.106Resident Self Testsp.107Availabilityp.107Logging Onp.107User Interfacep.108Information Typesp.109System Informationp.109Hardware Informationp.110Testsp.111Available Testsp.111Testing Optionsp.112Test Resultsp.113Service Logp.113Launching the RST Applicationp.113Closing the RST Applicationp.114Running a Basic Testp.114Running an Extended Testp.115Test Modulesp.117Running an Interactive Testp.118Viewing Error Logsp.119Ultrasound Application Logp.119Clearing the Ultrasound Application Logp.120RST Test Logp.121RST Log Codesp.122Clearing the RST Test Logp.122Test Result Detailsp.123Viewing Test Detailsp.123System Logp.124Clearing the System Logp.124Using the Service Logp.125Making an Entry in the Service Logp.125Modifying an Entry in the Service Logp.1267 Adjustmentsp.127Monitor Adjustmentp.127OSD Main Menup.127Opening the OSD Main Menup.127Navigating the OSD Menusp.128Adjusting VGA Vertical Sizep.128Adjust the VGA Vertical Sizep.128Changing the Boot Sequencep.130Change the Boot Sequencep.1308 Preventive Maintenancep.131Cleaning the Air Filterp.131Cleaning the Trackballp.132Cleaning the e-boxp.132Degaussing the Monitorp.1339 Troubleshootingp.134PC Won’t Bootp.134PC Boots, but Monitor is Blankp.135PC Boots, But Peripherals Don’t Workp.135Printer Problemsp.136HP940 Startupp.136Printer Installation Problemsp.136Electromagnetic Interferencep.13710 Service Proceduresp.138Overviewp.138Repair Philosophyp.139ESD Safeguardsp.139General Precautionsp.139Service Accessp.140Servicing Components Above the Lower Enclosurep.140Servicing Components in the Lower Enclosurep.140Rear Service Doorsp.141Front Service Panelp.141Side Panelsp.141Record Keepingp.141PC Servicep.142Software Service Proceduresp.142System Softwarep.142Reloading or Upgrading System Softwarep.142Backing Up and Restoring Presetsp.143Backing Up Customer Presetsp.143Restoring Customer Presetsp.146PC Componentsp.147Opening the PC Side Coverp.148Removing the PC from the Systemp.150Eboxp.151Replacing Ebox Circuit Boardsp.151Replacing the Motherboardp.152Replacing the Distribution Boardp.153Power Supplyp.154Removing the Power Supply and the Eboxp.154Separating the Ebox and the Power Supplyp.155Transducer Connector Modulesp.156Replacing Transducer Connector Modulesp.156Fanp.156Replacing the Fanp.156Monitorp.157Replacing the Monitorp.157Monitor Twivelp.158Disassembling the Twivelp.158System Control Panelp.158Replacing the System Control Panelp.159Monitor Support and Front Panelp.160Replacing the Monitor Support and its Front Panelp.160Cablesp.160System Control Panel and Monitor Cablesp.161Replacing the Upper Assembly Cable Bundlep.161Cartp.163Replacing Cart Wheels (Caster Assemblies)p.163Replacing the Top Panel of the Lower Enclosurep.164Lift Columnp.165Preparationp.165Preparing the System for Servicing the Lift Columnp.165Gas Springp.166Replacing the Gas Springp.166Mechlok and Cablep.167Replacing the Mechlok and Cablep.167Slidesp.169Replacing the Slidesp.16911 Cablingp.17112 Configurationp.174Disk Drive Settingsp.174Hard Drive Jumpersp.174CD Drive Jumpersp.175MO Drive Jumpersp.175VCR Switch Settingsp.175Sony B/W Printer Dip Switch Settingsp.17613 Partsp.177Overviewp.177Ordering Partsp.177New Partsp.178Exchange Partsp.178System Serial Number and Revision Levelp.178Major Assembliesp.178Transducer Connector Modulesp.179Internal PCp.180e-box and Power Supplyp.183Lower Assemblyp.185Lift Column Assemblyp.190Upper Assemblyp.193Top Shelfp.196VCR Option Partsp.198Softwarep.198Physio Option Partsp.199Exchange Part Numbersp.200Optional Peripheralsp.201Supplies and Accessoriesp.20214 Transducersp.20515 Glossaryp.206
Sources

References & links

Pages referenced in this manual — manufacturer, support and downloads.

Ask about the Philips Patient Monitor (M2540)

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the Philips Patient Monitor (M2540)

Be the first to ask — other owners or our team will answer.

Original document

Philips Patient Monitor (M2540) manual PDF

220 pages·4.4 MB·Audio System·PDF
Click any page above to open it here
Page 1 / …
100%
Loading pages…