system, and describes ESD precautions to be taken when servicing the
scan converter sections. This section also includes explanations of the
Section 1, “General Information,” presents a product overview and a
Section 3, “Safety,” discusses safety issues pertinent to the ultrasound
procedures that apply to the M2540A, including electrical safety tests.
safety, and system safety. A procedure for verifying complete system
This manual provides information for troubleshooting and servicing
Section 4, “Theory of Operation,” includes a technical overview of
system functions. Functional descriptions are given for each system
circuit board. Functional block diagrams show the scanner and the
theories behind the electrical safety tests in the Performance Tests
system, how to install peripheral devices, and how to configure the
description of major features, and lists applicable safety standards.
Section 5, “Installation,” explains how to unpack and assemble the
Section 6, “Performance Tests,” comprises all tests and diagnostic
the Philips M2540A ultrasound system. Brief overviews for each
The chapter emphasizes patient, operator, and service personnel
Section 2, “Specifications,” describes the physical and electrical
section in this book are listed below:
specifications of the M2540A.
system’s parameters.
operation is included.
system.
section.
Manual

Copyright © 2002 by Philips Medical Systems All rights reserved
M2540-92000-01 A November 2002


precautions, component removal and replacement, and peripherals installation. The chapter also
presents software procedures for backup and retrieval of customer presets, and for reloading or
Section 9, “Troubleshooting,” contains information and procedures for finding and repairing the
Section 10, “Service Procedures,” includes all service procedures, including a discussion of ESD
causes of faults in the M2540A, including software error codes. The error-code section opens
Section 12, “Configuration,” lists the jumper and switch settings for the M2540A’s disk drives.
Section 15, “Glossary,” is a glossary of terms used in this manual and in the ultrasound imaging
Section 8, “Preventive Maintenance,” includes procedures for periodic maintenance of the
Audience This manual supports the field service maintenance and repair of the M2540A Ultrasound
System. The user of this document is a qualified ultrasound electronics technician who has
Section 11, “Cabling,” includes tables listing all the standard and optional M2540A cables.
Section 13, “Parts,” comprises parts lists and exploded diagrams of the M2540A and its
Section 14, “Transducers,” lists the transducers compatible with the M2540A and their
Section 7, “Adjustments,” contains instructions for adjusting the system monitor.
completed training classes on the system and its peripherals.
assemblies. Parts ordering methods are discussed.
with a key for interpreting the codes.
upgrading system software.
characteristics.
system.
field.

If you have questions about the service manual, or if you discover an error in the manual, contact
Acrobat Reader. A list of bookmarks functions as a table of contents. Those bookmarks and
Format This manual is in Portable Document Format (PDF), for viewing on a laptop computer using
cross-references use hypertext links to provide access to the referenced information.
Conventions The following conventions are used in this manual:
Philips Ultrasound Technical Publications:
imply a sequential procedure.
to ensure reliable results.
are on the display.
Questions or
Comments

Non-Philips Medical Systems product names may be trademarks or registered trademarks of their respective owners.
Assistance Various support locations around the world can provide customers with technical assistance
regarding the ultrasound system. Customers should contact the sales office where they
purchased the system or the nearest Philips Ultrasound office for assistance.
To find your local service center phone number, go to:
Philips Medizin Systeme Boeblingen GmbH
3000 Minuteman Road Andover, Massachusetts 01810-1099
(425) 487-7000 or (800) 426-2670
Hewlett-Packard-Str. 2
P.O. Box 3003
Customer
Bothell, WA 98041-3003 · 71034 Boeblingen Germany · (978) 687-1501

This system complies with the
0123 marking is for Council Directive
Medical Device Directive.
Representative:
93/42/EEC.
The cautions explain potential
other symbols appearing on the
the user documentation for that
system complies with the FDA
this device to use by or on the
This symbol on the
This symbol on the
The United States Food and
Symbols on the System
electrical shock.
The monitor used in this
applicable at the date of
damage to equipment.
system or its probes:
part of the system.
Subcategory J).
statement:
Text:
only supplied power cords and
Do not operate the system in the
transducer for the first time, be
and Use” section of the chapter
explosion hazard, the safety of
Do not remove system covers.
To avoid electrical shock, use
grounded (3-hole) wall outlets.
sure to read the “Description
transducer. Also, for TEE and
review the “Electrical Safety”
guidelines for setting controls
that affect acoustic output and
Pay special attention to the
guidelines for fetal use, and
Before you use a specialty
sections in those chapters.
The warnings explain the
connect only to properly
that is applicable to your
measurements.
anesthetics.
Text:
© 2002 Philips North America
Reproduction in whole or part
is prohibited without the prior
written consent of the copyright
Philips Medical Systems makes
including, but not limited to, the
merchantability and fitness for a
Philips Medical Systems shall
damages in connection with the
Windows XP is a trademark of
The information contained in
no warranty of any kind with
this document is subject to
change without notice.
regard to this material,
not be liable for errors
contained herein or for
Cannon is a registered
All Rights Reserved.
or use of this material.
Printed in U.S.A.
particular purpose.
Corporation.
holder.

The M2540A ultrasound system comprises a cart, a system control panel, and a monitor. A foot
pedal on each caster locks and unlocks the front cart wheels. All four wheels swivel. The system
control panel and the monitor adjust up and down and swivel as a unit, and the monitor tilts and
Most peripheral devices mount on top of the lower enclosure. The optional plain-paper printer
The Resident Self Test (RST) software verifies system performance and helps diagnose problems.
foot switch, and a magneto-optical disk (MOD) drive. The VCR is available in either NTSC or
PAL configuration. For a complete list of optional components, see “Optional Peripherals” on
Optional peripheral components include a video cassette recorder, various types of printer, a
The cart’s lower enclosure contains the M2540A’s computer, disk drives, and ultrasound
is not mounted on or powered by the system, but is connected from a remote location.
CD-RW and floppy disk drives are standard equipment.
This chapter includes the following sections:
generating and processing boards.
swivels on its mount.
page 201.
Product

include important features of the cart, monitor, system control panel, Physio module, e-box, and
The panel is at the center of the lower part of the cart, and is accessible to the system operator.
Four wheels at the cart base provide system maneuverability and braking. All four wheels swivel,
and unlock the front wheels to prevent the cart from rolling. Locking the wheels immobilizes the
The system control panel includes four numbered option buttons whose functions are assigned
Major ultrasound system components are described on the following pages. These descriptions
References to the left and the right sides of the system are as viewed from the front of the cart.
A height adjustment lets you adjust the system control panel and monitor for operator comfort.
communication between the system and a network, and for controlling certain legacy printers.
The foot switch port, an equipotential lug, and the system ground lug are also on the I/O panel.
and the front wheels can be fixed straight and locked. Foot pedals on the two front wheels lock
components as standard equipment. These include the acoustic processor input/output (APIO)
Cart The cart supports the system and acts as the chassis into which all system components are
PC The M2540A uses a personal computer (PC) as a central processor. The PC houses several
by the customer. See the online Help file for details on configuring these controls.
An Input/Output (I/O) panel at the rear of the cart contains three connectors for
See “I/O Panel” on page 45 for more information about the I/O panel.
board, a video card, a CD-RW drive, and a floppy disk drive.
system during patient procedures.
installed and interconnected.
system power supply.
Description
System

magneto-optical disk (MOD) drive can also be installed in one of the drive bays. The VCR option
Panel The system control panel at the top front of the cart is a replaceable, self-contained module. See
“Replacing the System Control Panel” on page 159. The panel interfaces with the imaging system
through a USB cable that connects to the internal PC. Power for the trackball is provided via the
USB cable. All the other control-panel functions are powered by a cable from the system power
See “Internal PC” on page 39 for more detailed information about the PC and its components.
e-box The e-box houses the scanner circuit boards, and is accessible by opening the door at the right
System Monitor The monitor at the top of the cart is a 15-inch color display mounted on a “twivel” assembly.
includes three boards that reside in the PC’s PCI slots, and a VCR that mounts on the cart.
The system control panel contains a backlit alphanumeric keyboard, slide controls, rotary
The optional Physio module installs in one of the PC’s front drive bays. An optional
The twivel allows tilt and swivel positioning of the display for ease of viewing.
controls, hard-coded and software-driven keys, and a trackball.
There are several circuit boards in the e-box:
rear of the cart.
supply.

e-box at the right side of the cart. It accepts AC input sources from 100 V to 240 V, at 50 Hz or
120 VAC outlets for powering the monitor and two peripheral devices are at the rear of the power supply. An unswitched outlet powers the M2540A’s internal PC. When the system is shut
down, only the 120 VAC outlet to the PC is enabled. A green LED on the left front of the power
Physio Module The optional Physio module (see “Physio Module Option” on page 43) installs in one of the PC
An externally generated auxiliary analog input signal using the phone jack can substitute for the
3-lead ECG physio connector at the front panel. You can also input an auxiliary ECG waveform from an external patient monitor and display it, in all modes. The R-wave from this signal can
“Optional Peripherals” on page 201 lists all peripherals supported by the ultrasound system.
Power Supply The power supply is an enclosed, self-contained, replaceable module that mounts under the
The power supply provides all required AC and DC voltages to the system. Three switched
drive bays. The front panel of the module bears two input connectors:
The M2540A supports several types of optional peripheral devices:
Some peripherals mount on a shelf below the system control panel.
supply lights when the power supply is energized.
serve as a time reference for the system.
60 Hz.
Peripherals
Optional
Devices)
(VCR,
Image

The ultrasound system includes programmable presets that configure the system for best imaging
preferred configurations. If preset configurations are subsequently changed, they can be
quickly restored from the backup copy, without having to reset them manually. Keeping
For specific information regarding the configuration and use of presets or other features of the
results in a particular situation. Activating a preset initializes system settings to values that are
optimal for a chosen exam. The factory installed software includes several presets, tailored to
different applications. The user can define as many as 20 additional presets in each of 8 exam
a backup copy also eliminates the need to manually reconfigure the presets after a
types, to adjust system variables (including acoustic power) to any required configuration.
software upgrade. See “Backing Up and Restoring Presets” on page 143 for the
procedures to back up and restore presets.
system, see the online Help file.
Preset

Some of the keys found on a standard PC keyboard are not present on the M2540A. Many of the
keyboard. The M2540 keys act as PC-equivalent keys whenever the ultrasound application is not
running. The following table lists the missing PC keys and their equivalents on the M2540A:
functions performed by those keys are mapped to keys that are present on the M2540A
PC Key M2540A Key PC Key M2540A Key
F1 Patient F9 Blank 1
F2 Preset F10 Blank 2
F3 Review F11 Blank 3
F4 Report F12 Probe
F6 Help Delete Fusion
F8 Mic Page Down Right
F5 Setup Esc THI
F7 VCR Page Up Left
Table 1-1 Keyboard Equivalents
Keyboard

Ground wire leakage See Section 6, “Performance Tests”, and the Safety and Standards Guide
Wheel locks With wheel locks engaged, system remains stationary on slopes of up
External sharp edges The system exterior has no sharp edges, in compliance with IEC
Compliance For information about applicable safety standards and specifications, consult the Safety and
System tip over Will not tip over on an incline of up to 10° under normal use.
temperature System surfaces do not exceed temperature limits specified in
components The power supply and monitor comply with UL 2601-1 1997.
This section describes the safety specifications of the M2540A.
IEC 60601-1 and EN 60601-1
Power supply Complies with IEC60601-1
shipped with the M2540A.
to 5 ° in any orientation.
60601-1 and EN 60601-1
Standards Guide shipped with the M2540A.
Table 1-2 Safety Limits
Regulatory
Safety
Parameter Specification · Potentially hazardous · System surface

Specifications:
This section lists the specifications of the M2540A ultrasound system. The following specification
types are included in this section:

Depth (front to back) 103 cm (40.5 inches) See Figure 2-2
Width 53.5 cm (21 inches) See Figure 2-1
Height of lower enclosure 76.2 cm (30 inches) See Figure 2-2
Monitor to control panel front edge 43.18 cm (17 inches) See Figure 2-2
(front to back) 41.71 cm (16.42 inches) See Figure 2-2
Weight Less than 91 kg (200 lbs) including display but no peripherals See Figure 2-2
The tables and diagrams that follow detail the physical specifications of the M2540A.
Height (to monitor top) Lowest position: 129.5 cm (51 inches)
Table 2-1 Physical Specifications
Physical


129.5 cm (51 in) - 147.5 cm (58 in)
Figure 2-1 System Front Dimensions


Figure 2-2 System Side Dimensions


129.5 cm (51 in) - 147.5 cm (58 in)
Figure 2-3 System Rear Dimensions


Figure 2-4 System Top Dimensions

AC input 90 VAC to 264 VAC, 47 Hz to 63 Hz Ground impedance 200 milliohm maximum Dielectric withstand 1500 VAC mains to safety ground 2000 V mains to AC secondaries 4000 V mains to DC secondaries Load 1150 VA maximum. AC output 120 VAC, 60 Hz, quasi square-wave, 500 VA maximum
Table 2-2 Electrical Specifications
75 Hz refresh rate RGB color display
Screen size 15-inch diagonal Display format VGA, 800x600
Table 2-3 Main Display
Features Tilt ±30 degrees
Specifications
Electrical
Swivel ±135 degrees · Parameter Specification · Parameter Specification

The following table lists the specifications of the transducers that the M2540A supports.
PA 4-2 biopsy capable 21422A 2 to 4 MHz Advance Vision
CA 5-2 21425A 2 to 5 MHz Advance Vision
S4 1 21330A Cannon
c3540 biopsy capable 1 21321A 3.5 MHz Cannon
EC6509 endocavity 21336A Cannon
L7535 linear 21359A 7.5 MHz Cannon
TEE (Omni II) 21369A 6 MHz Cannon
L1038A small parts 21376A Cannon
15-6L intraoperative 21390A Cannon
D1914c 21221B 1.9 MHz Pencil
Transducer

Monitor 15-pin D connector on graphics adapter board Foot switch 9-pin female D-sub connector on the APIO board that carries foot switch signals from the external I/O panel
The following table lists the ports on the rear of the internal PC and describes their locations.
Speakers Green 3.5 mm stereo phone jack to system speaker amplifier
The following tables list the communication and connector specifications of the M2540A.
(optional) 1 Female phono (RCA) on TV and video converter board
S-Video output (optional) 1 4-pin mini circular DIN on SVGA to TV video card S-Video input (optional) 1 4-pin mini circular DIN on video capture card Print trigger output 3.5-mm phone jack on APIO board VCR audio output (line out)
(optional) 1 Green 3.5-mm stereo phone jack on sound card
VCR audio input (line in) 1 Blue 3.5-mm stereo phone jack on sound card
Microphone Pink 3.5-mm stereo phone jack
Table 2-5 PC Ports
and Commu-
Specifications
Connection
System I/O
nication
Port Specification · Composite video output

Com 1 9-pin D connector, RS-232; spare serial port used for serial VCRs
USB USB (A) for data and power to system control panel
Token ring 50-pin connector on the APIO board LAN RJ-45 to I/O panel for network communication
USB For black-and-white thermal printer
USB For remote plain-paper printer
Keyboard DIN circular connector
Mouse DIN circular connector
Table 2-5 PC Ports (Continued)
USB For color printer
PS-2 keyboard Not used
PS-2 mouse Not used
Parallel port Not used
Audio input on main board Not used
Port Specification

Each signal is a contact closure to ground, active low, and
Foot switch D-sub 9 Control signal from foot switch to APIO board
Composite video BNC Composite video signal from SVGA board
I/O Panel Ports The following table lists the ports on the I/O panel and the signals they carry.
LAN RJ-45 Communication with local area network
The following tables list the audio and video specifications of the M2540A.
The following table lists the specifications of the M2540A’s physio ports.
Print BNC Trigger signal from APIO board
Table 2-7 Physio Port Specifications
ECG 3 patient leads with R-wave detection Monitoring quality only
Table 2-6 I/O Panel Ports
Audio/Video
Specifications
Physio Port
TTL compatible · Port Connector Signal · Port Specification

Doppler spectrum away from transducer — right speaker
1.0 V p to p ±5% into 75 ohms (only with VCR option) Print Trigger 1 signal from the APIO board
Doppler spectrum toward transducer — left speaker
Microphone Faces front. 10 Hz to 20 KHz electret type
Table 2-9 External Video Specifications
Color Composite NTSC 3.58 (USA), PAL 4.43 (Europe)
Table 2-8 Audio Specifications
Response 130 Hz to 12 KHz
Audio

VCR 1.0 Vpp luminance incorporating sync and 0.3 Vpp chrominance as shown in Table 2-9. This timing is essentially RS-170 (60 Hz) or CCIR
Table 2-10 Video Output Specifications
monitor 0.7 Vpp RGB with TTL sync
(50Hz) video timing.
Specifications
Output Specification · RGBSync to system · S-Video Output to

Safety:
This section provides an overview of safety considerations for the ultrasound system. These
safety concerns apply to patients, operators, and service engineers. For more detailed safety
information, see the Philips Safety and Standards Guide.
The following topics are included in this section:

diagnostic image with minimal power output. A display standard presents this index using one of
For additional information on acoustical power settings and power index, see the Philips Safety
Acoustic output, expressed as an index, is displayed on the screen to allow the best possible
The index setting on the System Setup menu selects the power index used. The displayed
Although no harmful effects have been demonstrated for any of the ultrasound frequencies,
intensities, and exposure times used in examinations with Philips ultrasound systems, Philips
recommends that you consider the following, and use the lowest ultrasound exposure that
index is based on this setting and on preset configuration and imaging mode.
produces diagnostically acceptable information:
that affects transmit power level.
the following four power indices:
for display at any time.
and Standards Guide.
(Acoustical)
Transmit
Acoustic
Exposure
Power

For more detailed information on acoustic exposure, see the Output Display Standards and ODS
type of index displayed and the value of that index for the acoustic output currently being used.
The thermal index for bone (TIB) is used in applications such as second or third trimester fetal
Institute of Ultrasound in Medicine and the National Electrical Manufacturers Association, the
Philips ultrasound system displays power output indices related to the potential for bioeffects.
For example, if the output corresponds to a mechanical index of 0.8, the following is displayed:
The displayed index is one of four types: MI, TIS, TIB, or TIC. These ultrasound abbreviations
Real-time information related to the power output is displayed on the monitor, indicating the
In compliance with the Output Display Standard (ODS) jointly proposed by the American
Soft tissue thermal index (TIS) is used in cardiac, fetal, and abdominal scanning.
The cranial bone thermal index (TIC) is used for transcranial imaging.
scanning and neonatal cephalic (through the fontanelle) scanning.
Then use the imaging controls to fine-tune image resolution.
conform to the AIUM/NEMA Output Display Standard.
Acoustic Tables booklet.
AIUM/NEMA
Standard
Output
Display

If none of the conditions in “When MI is Displayed With the Normal Setting” on page 29 exist, if
power indices is selectable for display at any time, using the power index setting in the
index setting of Normal from the System Setup menu. This directs the system to choose an
Index Selection For automatic selection of a power index based on system mode, the user selects the power
transducer type, the imaging mode, and the selected power index. Any of the four
no transcranial preset is active, and any of the following conditions exist, TIS displays:
When TIS is Displayed With Normal Setting
MI is displayed if any of the following conditions exist:
index based on the active preset and imaging mode.
System Setup menu.

If none of the conditions in “When MI is Displayed With the Normal Setting” on page 29 exist, a
The displayed index value does not provide an exact value of the potential for adverse bioeffects
in the patient. However, for any patient, the higher the value, the higher the potential for adverse
bioeffects. The user can minimize the potential for bioeffects by keeping the index value as low as
possible, by choosing the right transducer and making adjustments. Minimizing examination time
preferred acoustic power display format, regardless of the mode, the transducer, or the preset
If the user selects an ODS setting other than Normal, the selected index type is used as the
likely that a particular patient's actual conditions are better than indicated by the index. The
transcranial preset is active, and any of the following conditions exist, TIS displays:
operator should be aware of patient conditions that mitigate the actual exposure.
Observe the following practices to avoid explosive hazards:
also minimizes bioeffects.
that is selected.
Hazards

power outlet on the imaging system, or from an isolation transformer that meets medical safety
standards. The 120 VAC outlets on the power supply are isolated. The specific peripherals listed
off. Be sure system power is on before the operation starts, and leave it on for the duration
peripherals within six feet of a patient unless the peripherals receive power from an isolated
Accidental contact with electrical circuits inside the system could cause serious injury.
in “Optional Peripherals” on page 201 meet medical device standards when installed in the
requirements, but do not meet medical device standards. Therefore, do not use system
Peripherals (such as a VCR or a printer) typically meet general electrical safety usage
Observe the following precautions to prevent electric shock:
grounded electrical outlets.
system as recommended.
of the procedure.
Peripheral
Electrical
Warnings

regulation dealing with levels of acceptable glutaraldehyde exposure in the working environment.
the system, since the ability to move any ultrasound machine is directly related to an individual’s
Philips does not sell glutaraldehyde-based disinfectants with its products. However, this type of
The ultrasound system has been designed to be as lightweight and mobile as possible. However,
approximately 100 kg (220 pounds). Because of this weight, you must use caution when moving
injured their backs moving similar systems. These complaints could not be directly tied to one
particular incident. They do, however, point out the need to be careful in transporting medical
To minimize exposure to glutaraldehyde fumes, make sure the area is well ventilated and use
Although the system is designed to be mobile, remember that it is very heavy, and that you
Some sonographers, particularly those weighing less than 100 pounds, have stated that they
The United States Occupational Safety and Health Administration (OSHA) has issued a
disinfectant is recommended for disinfection of transesophageal (TEE) or endocavity
the system weight—including the weight of the monitor, a printer, and a +VCR—is
tem, disconnect the system power cord, and disconnect all external devices.
the System To move the system, take the precautions listed in the following sections, and
must take precautions when moving the system.
equipment such as an ultrasound system.
system and put them in a packing box.
appropriate eye and skin protection.
size and strength.
transducers.
Moving the
Glutaral-
Exposure
dehyde
System

Do not move the system over uneven elevator entrances by lifting the machine or any part of the
feet or steeper than 5 degrees. (Wheelchair ramps are usually less than 5 degrees.) Avoid ramps
Always use two people when moving the ultrasound system up and down ramps longer than 20
protocol, the system will not tip over on an incline of up to 10 degrees in any direction. When
System Tilting The system has been tested for stability using the IEC 60601-1 test protocol. Following this
Always use the handle at the front of the cart to move the system from place to place.
that are steeper than 10 degrees, to eliminate the danger of the system tipping over.
this amount of incline is exceeded, there is a potential for the system to tip over.
the System This system is equipped with a front handle and brakes on the front wheels.
machine.
When Moving
Moving on
Ramps or

The system’s weight on an extended loading lift may cause the transporting vehicle to tilt, which
Be sure that the transporting vehicle can handle the weight of the system (or systems) plus the
chocks, restraining straps, or other similar types of constraints as an added safety measure. Do
Always engage the wheel locks while transporting the system in a vehicle, and use restraining
Always secure the ultrasound system while it is on the loading lift so that it cannot roll. Make
sure the control top is locked, to prevent its pivoting. Engage the wheel locks and use wood
straps to secure the system in place. Do not rely on the wheel locks to hold the system on
Be sure the load capacity of the loading lift can accommodate the weight of the ultrasound
Load and unload the ultrasound system while the transporting vehicle is parked on a level
surface. The system’s weight can easily cause it to roll on any incline.
system. A minimum capacity of 550 pounds is recommended.
can cause the system to exceed 10 degrees of incline.
not attempt to hold the system in place manually.
could cause personal injury or system damage.
inclines greater than 5 degrees.
passengers.
Transporting
Vehicle

Electrotechnical Commission (IEC 60601-1-2). This IEC standard has been adopted in Europe as
Medical devices can generate or receive electromagnetic interference (EMI). The EMC standards
susceptible to EMI generated by other RF energy sources. Examples of other sources of EMI are
medical devices, information technology products, and radio and television transmission towers.
describe tests for both emitted and received interference. Emission tests deal with interference
Ultrasound systems are designed to receive radio frequency (RF) energy and are therefore
Only a physician can determine if an artifact caused by radiated interference has a negative
Be sure the ultrasound system is firmly secured while inside the transporting vehicle. Any
international standard for EMC with medical devices, as determined by the International
movement, combined with the system’s weight, could cause the system to break loose.
This system has been tested for electromagnetic compatibility (EMC) according to the
generated by the device being tested. The Philips ultrasound system does not generate
See “Electromagnetic Interference” on page 137 to identify sources of EMI.
Tracing the source of radiated interference can be a difficult task.
transporting precautions listed in the sections above.
interference based on the tests described in the standards.
impact on image quality and the subsequent diagnosis.
the European Norm EN 60601-1-2.
exceed the lift's load capacity.
Electromag-
Interference
Avoiding
netic
netic

first sign of an artifact is seen. This method has two advantages: It removes the issue of subjective
The criteria used for measuring degradation are not specified by the standard, and can vary with
Discharge Electrostatic discharges can cause the ECG heart rate display to increase by 10 to 15% for a few
Immunity is defined in the standard as the ability of a system to perform without degradation in
seconds after the discharge. However, the ECG heart rate display returns to normal within four
assessment that can be subjective. The simplest way to assess degradation is to note when the
receive and amplify low-level signals in the same bandwidth as the interference to which it is
frequencies. This testing showed PW Doppler to be the most susceptible to radio-frequency
the presence of an electromagnetic disturbance. Degradation in image quality is a qualitative
The M2540A ultrasound system is subject to certain restrictions that are described in the
electrostatic discharge immunity levels for their systems. This type of device is designed to
Philips has tested each class of transducer for every operating mode over a wide range of
Please see the Philips Safety and Standards Guide shipped with your system for additional
Test Results The EMC standards require that manufacturers of patient-coupled equipment specify
decision-making and provides the most stringent test results.
information about compliance with EMC standards.
following sections.
the manufacturer.
interference.
susceptible.
seconds.
Restrictions
Electrostatic
for Use

electromagnetic fields or currents into patients. Because imaging ultrasound frequencies are also
in the radio frequency range, ultrasound transducer circuits are susceptible to radio frequency
Electrosurgical units (ESUs) and other devices intentionally introduce radio frequency
interference. While an ESU is in use, the noise generated severely interferes with the
black-and-white image and completely obliterates the color image.
gical Units

The M2540A’s internal PC replaces the back-end electronics used in earlier ultrasound systems
to handle display processing. The beam processor/acoustic processor (BPAP) board handles the
Overview This section provides a technical overview of system functions. The discussion focuses on the
main functions and features of the system’s PC, circuit boards, and power distribution. Main
system functions are also shown in the functional block diagrams at the end of this section.
This section covers the following areas of the system:
acoustic and beam processing front-end functions.

Internal PC The M2540A uses an internal PC to coordinate and perform many of the functions that required
dedicated circuit boards in earlier ultrasound systems. The power button on the front of the PC
is the system power switch. Turning on the PC starts the ultrasound system.
Components The M2540A’s internal PC contains the following standard components:
The PC motherboard’s ports are used as follows:
Standard PC

The VCR option includes several cards inside the PC and an adapter that connects to the SVGA
communicates with the ultrasound system through a USB cable to the PC motherboard. See
tor powers the module, but the module does not use the drive bay’s data port. The module
“Physio Module Option” on page 43, and Figure 4-3 on page 51 for detailed information on
TR Boards The M2540A’s e-box contains two transmit-and-receive (TR) boards. Each has a connector that
plugs into the motherboard and another connector that passes through the motherboard and
representation of how the VCR option’s components function in the ultrasound system.
See “VCR Option Parts” on page 198 for a list of all the components of the VCR option.
output of the PC’s graphics adapter board. See Figure 4-1 on page 48 for a detailed
The following cards are part of the VCR option, and reside in the PC’s PCI slots:
removable disks. The MOD uses the drive bay’s data and power ports.
The PC’s drive bays accept the following components:
Optional Drive-Bay Components
the AGP slot) to S-Video for the VCR
and to the monitor
the Physio module.

input bus. The last TR board in this chain outputs the final summed radio-frequency (RF) data to
are daisy-chained together to sum the outputs. The RF data output of one TR board is the input
to the next TR board, and the outputs of the TR boards are summed to form the received beam.
and times each of the analog drivers. The delay generators receive the appropriate delay values
The received signal is preamplified, TGC amplified, low-pass filtered, and digitized (in all modes
After being digitized, the signal on each line is delayed appropriately. The outputs of all channels
filters, 10-bit A/D converters, and one eighth of a digital beam-forming circuit. Each TR board
The BPAP board generates the timing to start and stop each firing line, and sends the transmit
transducer. Analog drivers generate the transmitted RF data. A digital timing generator delays
except CW). The BPAP board generates TGC control signals through the demodulator board.
plugs into the distribution board. The two boards form electronic beams of the transmit and
sends on its output bus a value that is the sum of the data from its 32 channels plus its 18-bit
The transmit and receive beam-forming coefficients for acoustic lines are downloaded on the
receive signals. Each TR board contains 64 transmitters and 32 preamps multiplexed into 64
receive channels. This allows multiplexing the system’s 64 active channels into 128 element
arrays. Each receive channel consists of time gain compensation (TGC) amplifiers, low pass
The CW Doppler signal is mixed to an analog intermediate frequency (IF) signal. The CW
signal to the TR board, through the distribution board and the connector modules to the
from the BPAP for each channel. The system power supply controls the amplitude of the
Doppler output is a single differential signal routed to the demodulator board.
The summed output signal (RF DATA) goes to the demodulator board.
coefficient bus from the BPAP board.
the demodulator board.
transmitted waveform.

motor. It also measures the position of the angle-position sensor. The signal is digitized and sent
BPAP Board The beam processor/acoustic processor (BPAP) board contains two processors. One of them
Circuitry The Omni Probe control circuitry on the BPAP board supplies power to the TEE transducer's
The beam processor circuitry monitors thermistors in the transducer and insures that high
controls beam-forming, line-to-line operation. The other handles the processing of scan
voltage is not applied to the transmitter unless the thermistor values are within limits.
motor control and temperature monitoring)
to the APIO board for monitoring by the PC.
conversion, color flow, and Doppler signals.
The BPAP board does the following:
sends them to the TR boards
that can be connected

and CW phased and non-imaging Doppler modes, and color flow mode. It sends a baseband data
and the e-box. The APIO board also handles input from the foot switch’s three control buttons
The demodulator board performs image and Doppler demodulation for BMode, MMode, PW
APIO Board The acoustic processor I/O (APIO) board is the token ring interface between the system’s PC
(via the I/O panel) and forwards them to the PC. The board also generates trigger signals for
The self-contained Physio module resides in one of the PC’s drive bays. The module receives
power from the drive bay’s power connector, but it communicates with the system through a
electromagnetic interference. See Figure 4-3 on page 51 for a detailed representation of the
USB cable to the PC’s motherboard. A shielded and grounded enclosure prevents
The demodulator board contains the following:
The board handles these signals from the PC:
module’s functions and connections.
stream to the BPAP board.
legacy printers.
Demodulator
Option
Board

This board connects all data paths between the TR boards, the BPAP board, and the distribution
The distribution board passes transducer signal inputs to the imaging system. The system power
board. It is also the distribution point for DC power and ground to all boards (the BPAP board
and distribution board also receive power directly from the system power supply.). It also sends
supply provides high voltage directly to the distribution board to fire transducers.
Connectors The Physio module has two connectors on its front panel:
Functions The module’s primary functions are as follows:
Signals that pass through this board include
filtered high voltage to the TR boards.
Distribution
Connector
Board and
Modules
System
M2540A’s PC · modes · ducer

rotary and slide controls, and the trackball. A USB cable carries control inputs from the control
I/O Panel The I/O panel bears four connectors and two ground points. The main system ground used for
The system control panel contains the circuitry to process control inputs, the system keyboard,
entire assembly, including circuit boards, bezel assembly, and labels, is available as a replaceable
panel to the PC. An auxiliary cable carries power from the system power supply. The system
The four transducer module bays accept the following types of modules in any combination:
Circuit boards inside the system control panel are not field serviceable or replaceable. The
All control inputs and LEDs are controlled by the system’s PC, and their functions are
control panel circuit board also carries the amplifiers for the system’s stereo speakers.
safety testing, and an equipotential post are the two ground points on the panel.
See Table 13-2 on page 179 for a list of the modules and their part numbers.
determined by resident control-panel and system software.
The I/O panel’s connectors are
Transducer
Connector
System

Power Supply The M2540A power supply is a self-contained unit that accepts AC input from 90 to 264 V, at 50
also provides all DC power required by system components. No switch settings are required to
configure the power supply for different AC mains; the power supply automatically adjusts to use
the power connected to it. Except for the 120 VAC to the PC, all power output is activated by a
signal from the PC. The system cooling fan is the only serviceable part on the power supply. See
Figure 4-1 on page 48 for details of the power supply’s outputs. The power supply activates when
it receives an “AWAKE_N” signal, which is generated by the PC, and then sent through the BPAP
part. Rotary and sliding knobs, rotary encoders for the knobs, the trackball assembly, and the
or 60 Hz, and supplies 120 VAC at 60 Hz to the system’s PC, monitor, and two peripherals. It
trackball cable are all available as separate parts.
The system power supply does the following:
board to the power supply.
board

gram does not include the e-box or the system’s power supply. Those components are
This section contains block diagrams of the system’s components and their connections.
Figure 4-1 shows the M2540A system with the VCR option installed.
The section includes the following block diagrams:
components of the ultrasound system.
shown in Figure 4-2 on page 49.
Functional
Diagrams
Diagram
System
Block


Figure 4-1 System Block Diagram with VCR Option
USB printer · To remote

Figure 4-2 shows the system’s power supply and the e-box and their connections to each other

To APIO board in PC
Figure 4-2 e-box and Power Supply Block Diagram
and other components.
Functional Block
Diagram
Supply
To system

The token ring cable carries the following signals between the BPAP board and the APIO board:
for the PC is always on)

Figure 4-3 shows how the Physio module functions, its internal connections, and connections to
EZ-USB
M
U
USB
0.35 to 30Hz
Figure 4-3 Physio Module Block Diagram
other components.
Diagram
Physio

This section provides background and supporting information for the transducer leakage tests in
The test for transthoracic and endocavity transducers differs from the test for TEE transducers
insulation and provides an electrical path between the submerged lead wire and the grounded
In transducer safety tests, a container filled with saline solution functions as a conductive
medium (see Figure 4-4 on page 53). The solution penetrates any faults in the transducer
Section 6, “Performance Tests”; this information is not required to perform the tests.
only by the extent that the transducers are submerged in the test solution.
inner transducer shield.
Transducer
and Theory
Test Setup

Closed for 2nd single fault condition

Open for 1st single fault condition
Case Leakage, Ground Conductor for ground wire leakage test Mode selector on safety analyzer:
Figure 4-4 Transducer Leakage Current Test Setup and Theory Diagram
Ground condition:
ECG for transducer leakage test
Closed for 1st single fault condition
Open for 2nd single fault condition
Closed for normal condition

Leakage current Ichassis, driven by the line supply, flows through the stray capacitance C between
Analyzer. Select ECG for transducer leakage and Case Leakage, or select
Impedance between the transducer’s metal parts and the test electrode:
through the ECG lead wire. (This is the mode selector on the Safety
S = Switch that connects the ammeter either directly to the chassis or
C = Stray capacitance from the system's power wiring to the chassis.
≈ 800 KΩ if the outer insulating layer is intact
< 500 Ω if the outer layer is compromised
Ground Conductor for chassis leakage.)
the patient: 100 cm for adult TEE, 60 cm for pediatric TEE.
A = Microammeter in the safety analyzer
being careful to not submerge the connector.
the primary wiring and the system’s metal chassis.
Table 4-1 Key for Figure 4-4
Imeasured = Leakage current
Insertion depth:

The leakage current normally flows from the chassis safely out through the ground wire. If there
is a fault in the transducer insulation, some of the current follows this path and is measured by
the safety analyzer. This Itransducer is still fairly low, unless the chassis is not properly grounded.
Figure 4-5 Transducer Leakage Current Test Diagram for the Normal Condition


When an open ground condition is imposed, all Ichassis is forced through the transducer. This
Figure 4-6 Transducer Leakage Current Test Diagram for 1st Single Fault Condition
Figure 4-7 Transducer Leakage Current Test Diagram for 2nd Single Fault Condition


Itransducer is still fairly low unless its sheath is compromised.

flows through the transducer. This Itransducer is still fairly low, unless the chassis is not properly
In the 1st single fault condition transducer leakage test (Figure 4-6 on page 56), all the leakage
current Ichassis flows through the analyzer by way of the transducer. In that condition, only the
higher potential for leakage current Ichassis. Most flows safely through the ground wire; some
With the transducer circuit disconnected, all of the leakage current Ichassis flows through the
Imposing an open neutral condition prevents all system current from flowing. This creates a
Figure 4-8 Ground Wire Leakage Test Diagram (for Comparison with 1st Single Fault
test is low when compared with Imeasured in the ground wire leakage test.

resistance or impedance of the respective paths varies.
analyzer by way of the ground wire.
grounded.

Transducer leakage current equaling 80% or more of ground wire leakage current indicates a
causing significant Itransducer, and yet that leakage current could still be within acceptable limits.
Sheath integrity is tested this way because there could be a break in the transducer sheath
Comparison to the ground wire leakage current is the only way to ensure sheath integrity.
fault or break in the transducer insulation.

Overview This chapter includes procedures for installing and setting up the M2540A.
Complete the following steps to assemble the M2540A ultrasound system:
Installation When installing peripherals, keep the following points in mind:
➤Installing the Power Cord and Monitor
The following sections are in this chapter:
Assemble the
Installation
Peripherals
System
Monitor

between the lower peripheral and the inside edge of cart top.
➤Installing a Sony Color Printer
➤Installing a VCR

➤Installing a Sony Black-and-White Printer
black-and-white printer.

Voltage When shipped from the factory, the monitor signal input voltage is set to 1.0 V. This must be
Complete the procedures in this section to set up the system for the customer’s use.
➤Setting the Monitor Signal Input Voltage
reset to 0.7 V, using the following procedure:
➤Powering On the System
System
Startup



The Advanced Controls menu displays.
The OSD Main Menu displays.
Figure 5-2 Advanced Controls Menu
Figure 5-1 OSD Main Menu

Administrator must be available to supply configuration data.
The following procedures configure the system’s software:
Preparation ➤Preparing the Sony Black-and-White Printer
➤Preparing the Sony Color Printer
roll.)
System


Localization ➤Setting the Localization Parameters
Figure 5-3 Setup Window


The Date and Time Properties window opens.
Figure 5-4 Top Border Window


Figure 5-5 Date and Time Properties Window

Figure 5-6 Regional and Languages Options Window (Regional Options)

dates in the display to be truncated.

online Help displays in, if that language is available. If it’s not, Help uses the same language as
Figure 5-7 Regional and Languages Options Window (Languages)

the rest of the system.

system. You may be prompted to restart or to “use existing files.” Click Yes in either
Figure 5-8 Regional and Languages Options Window (Advanced)

case.

Installation The M2540A leaves the factory with all purchased options enabled. If an option must be enabled

in the field, complete the following procedure:
Figure 5-9 Setup Window (Options)
➤Installing Options


The selected option is enabled.
Figure 5-10 Options Installation Window

Complete the following procedures to configure the system for DICOM communications over a
network. The procedure requires information from the customer’s network administrator.

➤Changing the System’s Network Parameters
The DICOM Setup window displays.
Figure 5-11 DICOM Setup Window
Network
DICOM

➤Using the Field Service Options Window for Advanced Network Settings
following procedure to change the network settings.

The Field Service Options window opens.
Figure 5-12 Setup Window (Options)

work as an ultrasound imaging device. This is true even if no settings are changed.
The display changes to Windows Classic format.

The Control Panel window opens.
Figure 5-13 Field Service Options Window


Figure 5-14 DICOM Setup Window (Servers & Roles)
➤Adding Server Information

The M2540A tests communication with the server, and opens a window that shows the
to identify the server in error messages and dialog boxes.
➤Assigning Servers to DICOM Roles
resolves to an IP address.
result.

The following procedure configures how the system automatically deletes studies.

Figure 5-15 DICOM Setup Window (Servers & Roles)
➤Configuring Autodelete


Figure 5-16 DICOM Setup (This System)


Figure 5-17 Disk Full Strategy Window
➤Configuring Logging


Figure 5-18 DICOM Setup Window (Diagnostics)

The following procedures configure the system for the installed peripherals and system

➤Assigning Record Buttons
Figure 5-19 Setup Window (Peripherals)
hardware.
Hardware
System

The Define Option Keys window opens.
➤Assigning Option Buttons
Figure 5-20 Setup window (Options)

The system is configured at the factory for NTSC format. To reconfigure the system for PAL

➤Configuring the System for PAL Format
format, complete the following procedure:
Figure 5-21 Define Option Keys Window
features.

The Install Software Drivers window opens.
Figure 5-22 Setup Window (Peripherals)

Cardiologists prefer a bluer tint to the monitor display than radiologists do. Pressing the two left
The drivers for Sony UP-D895, Sony UP-D21MD, HP920, and HP1200 printers are loaded
The driver is reinstalled with the correct configuration. This takes about 20 seconds.
automatically. These printers are ready to use when connected to a USB port.
monitor buttons together toggles between the two color settings.
The new setting takes effect after the system is restarted.
➤Setting the Monitor Color Temperature
To set the monitor color temperature:

Presets” on page 143). Make a backup floppy disk of the customer’s presets and configura-
tions, and store it behind the lower enclosure’s left rear door. If the software is reloaded
The customer needs some instruction in maintenance and use of the system:
later, the backup floppy disk lets you quickly restore settings and options.
explain that the queue is empty if all studies have transferred successfully.
the importance of the filter to system performance.
power cord.
page 132.)
Training

Overview This chapter includes all test procedures for diagnosing problems with the M2540A and assuring
The tests described on the following pages are part of a comprehensive preventive maintenance
between the chassis and ground. See “Chassis to Ground Resistance Test” on page 90.
program. They detect abnormalities that could prove dangerous to patient or operator.
further safety testing is needed when the power supply is installed or replaced.
leads. See “Ground Wire Leakage Current Test” on page 92.
The following safety tests are described in this section:
The following sections are in this chapter:
and functions are not compromised.
its safe and accurate operation.
Safety Tests
Electrical

applying mains voltage to the leads, and checking the transducer for leakage. See “Transducer
step in any repair or upgrade, to maintain agency approved status. The safety analyzer is
also useful in detecting abnormalities of line voltage, grounding, and total current loads.
the leads, to assure that the system is isolated from such voltages. See “ECG Lead Isolation
measurements can be performed with multifunction instruments like the HP 3469A multimeter.
The limits referenced in these tests are prescribed by UL, CSA, and IEC. Local regulations may
transducer, which is tested for leakage. See “Transducer Leakage Current Test Procedure”
module for electrical leakage. See “ECG Lead Leakage Current Test (Source)” on page 94.
connected to the ultrasound system. The system sends normal operating voltages to the
All tests can be performed using commercially available safety analyzer test equipment. Basic
Figure 6-1 on page 91 shows the basic electrical concept of the ground resistance test. Use a
This test checks the entire system for electrical resistance between the chassis and ground.
safety analyzer and complete the procedure that follows the figure.
Follow the instructions of the analyzer manufacturer.
Isolation Leakage Current Test” on page 102
Leakage Current Test” on page 96.
require additional tests.
on page 100.
Chassis to
Ground

10 amps. Test currents in excess of 1 amp will likely exceed the impedance limit specification, and can damage the RFI protection finger contacts. Exposed metal on the transducer assembly,

including the connector, are for RFI and are not safety grounds.
To test the system for resistance between chassis and ground:
➤Chassis to Ground Resistance Test Procedure
Figure 6-1 Chassis to Ground Resistance Test Diagram
R = 200 milliohms maximum
300e036 · (resistance)

shows the basic electrical concept of the test. Use a safety analyzer and complete the procedure
voltage drop across the entire length of the power cord and to the system chassis. The reading
A comprehensive ground impedance measurement can be performed using the safety analyzer.
This test checks the entire system for leakage on chassis ground wires. Figure 6-2 on page 93
The impedance test drives a load current through the ground wire while measuring the AC
ground connection is open, do not touch the M2540A chassis or the patient cable.
primary wiring.
that follows the figure.
will be in milliohms.
of the system.
Current Test
Leakage

simultaneously for Single Fault condition

1000 microamperes Single Fault condition 7ASW025
➤Ground Wire Leakage Current Test Procedure
Figure 6-2 Ground Wire Leakage Current Test Diagram
1000 microamperes Single Fault condition
for Normal condition in Figure 6-2 on page 93.
To test the system’s ground wiring for leakage:
internal PC.
(current)

Figure 6-3 on page 95 shows the basic electrical concept of the ECG Lead Leakage Current test.
A reading higher than acceptable limits can indicate a problem with the power cord, with its
This test checks ECG leads connected to the optional Physio module for electrical leakage.
ground connection is open, do not touch the M2540A chassis or the patient cable.
to the limit specified for Single Fault condition in Figure 6-2 on page 93.
Use a safety analyzer and complete the procedure that follows the figure.
associated connections, or with the power transformer.
ECG Lead
Leakage
(Source)


Neutral wire open (ground closed) for second Single Fault condition
➤ECG Lead Leakage Current Test Procedure
Figure 6-3 ECG Lead Leakage Current Test Diagram
50 microamperes Single Fault condition
I = 10 microamperes Normal condition
For

condition in Figure 6-3. Write down the highest value; this is the value for the first single fault
concept of the ECG Lead Isolation Leakage Current test. Use a safety analyzer and complete the
condition in Figure 6-3. Write down the highest value; this is the value for the second single
This test checks ECG leads by applying voltage from the mains to the ECG leads, to check that
Failure to meet the specified limits may indicate a fault with the isolation of the Physio module.
the system is isolated from such voltages. Figure 6-4 on page 97 shows the basic electrical
procedure that follows the figure.
fault condition.
condition.
page 95.
ECG Lead
Isolation
Leakage

line voltage. Do not touch the chassis or the ECG cable while performing the test. Keep the
ECG cable at least 20 cm from any grounded or conductive surfaces.

➤ECG Lead Isolation Leakage Current Test Procedure
Figure 6-4 ECG Lead Isolation Leakage Current Test Diagram
open neutral conditions.
Power cord connected to AC power.
ECG patient

patient safety, it is very important to verify the integrity of the insulating layers of all transducers.
When performing the safety tests, saline solution in a container is used as a conductive medium
about the transducer tests. This is supplemental information, and is not required to complete the
Tests Two electrical leakage current tests for transducers are described in this section. To insure
(see Figure 4-4 on page 53). The solution penetrates any cracks or holes in the transducer
insulation and provides an electrical path between the submerged lead wire and the inner
transducer is a TEE, check that its steering controls are working properly.
reading. Refer to the analyzer’s documentation for more information.
Figure 6-4 on page 97.
transducer shield.
tests.

system. The system sends normal operating voltages to the transducer, and leakage is measured
This test checks a transducer for electrical leakage while it is connected to the ultrasound
using a safety analyzer. Figure 6-5 on page 100 shows the basic electrical concept of the
Transducer Leakage Current test. Complete the procedure that follows the figure.
The following tools are required for transducer safety tests:

Neutral wire open (ground closed) for second Single Fault condition Ground wire open for first Single Fault condition, lead
500 microamperes Single Fault condition I = 100 microamperes Normal condition,

For type transducers:
➤Transducer Leakage Current Test Procedure
Figure 6-5 Transducer Leakage Current Test Diagram
50 microamperes Single Fault condition I = 10 microamperes Normal condition, For type transducers:
wire · Hot -

saline to contact its control handle. Liquid inside the control handle causes corrosion and failure.
page 100. (Match the transducer type symbol on the transducer connector or cable with the
body of the patient: 100 cm for adult TEE, 60 cm for pediatric TEE.
cable. Be careful not to submerge the connector.
symbol in the figure.)
transducer:

Values exceeding the limits may indicate a fault in the transducer housing or the cable sheath.
To locate a fault precisely, repeat the measurement while slowly inserting the transducer into
of the cable. Their leakage currents can measure within the specified limits even if there
Current test. The transducer is submerged in a container of saline solution. Mains AC voltage is
Figure 6-6 on page 103 shows the basic electrical concept of the Transducer Isolation Leakage
If the transducer leakage is greater than 80% of the chassis leakage, there is a fault in the
introduced into the saline solution, and leakage current in the transducer is measured using a
transducer. To locate the fault, slowly insert the transducer in the saline, as in step 10.
the saline. The aberrant reading appears when the fault in the cable enters the saline.
for transducer leakage current, Single Fault condition obtained in step 10.
is a fault in the outer skin. Step 11 tests for this condition.
safety analyzer. Complete the procedure that follows the figure.

contact with the line voltage. Do not touch the chassis, or the ECG or transducer cables while
performing the test. Keep the ECG and transducer cables at least 20 cm from any grounded or

➤Transducer Isolation Leakage Current Test Procedure
Figure 6-6 Transducer Isolation Leakage Current Test Diagram
open neutral conditions.
conductive surfaces.
wire · 1101e30

saline to contact its control handle. Liquid inside the control handle causes corrosion and failure.
body of the patient: 100 cm for adult TEE, 60 cm for pediatric TEE.
cable. Be careful not to submerge the connector.

PC problems can often be diagnosed using the IBM Drive Fitness test or Checkit Pro.
You must make a bootable floppy disk to use the Drive Fitness test.
➤Testing PC Functions Using the Drive Fitness Test
http://www.storage.ibm.com/hdd/technolo/dft/dftnew.htm
➤Creating a DFT Bootable Floppy Disk
Fitness Test The Drive Fitness test (DFT) is described at:
into a temporary directory on your PC.
PC

Checkit Pro Test PC functions using the CheckIt Pro software on the Diagnostic floppy disk.
http://service.boulder.ibm.com/storage/hddtech/DFT32-V300USER-DOC.pdf
first. (See “Changing the Boot Sequence” on page 130.)
➤Testing PC Functions Using Checkit Pro
For detailed information on using DFT, go to:

Customers or service personnel who log on with a valid user name and password can create and
The M2540A Resident Self Tests (RST) are a suite of tests that verify system hardware, using no
The stand-alone RST application provides access to all the tests. RST also provides information
The logon window also includes a check box that skips the logon dialog the next time the RST
The RST suite consists of tests for custom M2540A hardware and software. The Resident Self
Tests also exercise individual subsystems such as the acoustic processor and the beam former
information, test pass and fail results, and errors can be viewed, saved, and printed by the RST
application opens with a logon window, which asks you to enter a user name and a password.
Availability The RST application is available whenever the ultrasound application is running. You can also
application is launched. Only service personnel can enable that option, and it only applies for
Logging On To launch the RST application, see “Launching the RST Application” on page 113. The RST
launch RST as the system powers up. When it starts, the RST application stops the main
24 hours. After 24 hours, the logon dialog shows at launch, even if the option is selected.
on hardware and software, and service information on the ultrasound system. System
modify service log entries. See “Using the Service Log” on page 125.
If RST is installed on the system, logging on is not required.
external instrumentation.
ultrasound application.
application.
processor.
Tests

User Interface You can control the RST software using the system control panel’s keyboard and trackball. When
hierarchy of supported functions. A list view is on the right side. The list view displays information
using the system control panel, the two keys just above the trackball perform the functions of
mouse buttons. The Enter key acts as a left mouse button, and the Select key acts as a right
A tab view is on the left side of the RST window. On each tab is tree control that displays a

related to items selected in a tree control.
Figure 6-7 shows the RST window.
Figure 6-7 RST Window
mouse button.

log on the Log tab, for service personnel to record information about the system. See “Using
Choosing a tree item and pressing the Select key opens a drop-down list that contains Save Info
log entries in the service log. See “Viewing Error Logs” on page 119. There is also a service
and Print Info functions. Information can be saved to a text file or printed on a connected plain
institution name, installed software and hardware options, and installed peripherals. Figure 6-7,
The software information provided includes the revision number and the revision date and time
of the operating system, the main ultrasound application, and the common off-the-shelf (COTS)
Types The RST application provides information on the hardware and the software of the ultrasound
system. You can select the Info tab and choose specific information types in the tree control.
The RST application provides system information that includes the system serial number, the
There are three tabs in the RST window:
the Service Log” on page 125.
on page 108 illustrates this display.
paper printer.
page 109.
software.
page 111

The hardware information provided includes information about the custom M2540A boards and

transducer connector modules, identified as shown in the following figure:
connector modules installed.
Figure 6-8 Hardware Information Display

five minutes. The test requires no operator interaction while executing. See “Running a Basic
The following tests are available when you log on using a valid service name and password:

Test” on page 114 for instructions on running basic tests.
Tests The following figure shows the RST Test display.
Figure 6-9 RST Test Display

selected tests run until completed. Any errors occurring during the tests are recorded in the
operator interaction while it executes. The test verifies detailed system operation with fault
The left-hand panel under the RST Test tab contains three testing options and the Run button
participation. Included are tests of the display and system control panel. See “Running an
isolation to the board level. This test reports only failures that adversely impact system
performance. See “Running an Extended Test” on page 115 for instructions on running
that begins a selected test. When an interactive test begins, the Run button is grayed out.
Interactive Test” on page 118 for instructions on running interactive tests.
Disconnect all transducers before beginning any RST testing.
value of zero runs tests continuously.
The three test options are:
extended tests.
error log.

Test Results The test results are shown in the list view on the right-hand side of the RST window. The display
shows the name of each test, the components it tested, and the number of times the component
You can enter notes in the service log section of the event viewer, to record service procedures
To view details of the test results, click the Log tab, and select the information type from the
Event Viewer tree. For information on errors generated during tests, see “RST Test Log” on
You can launch the RST application as the PC boots up or while the ultrasound application is
you perform or problems you encounter. See “Using the Service Log” on page 125.
When it launches, the RST application closes the ultrasound application.
When the tests are finished, the following message box opens:
To launch the RST application when the system is fully booted:
To launch the RST application during system boot-up:
running. The procedure is the same in both cases.
➤Launching the RST Application
passed and failed the test.

page 121.

The basic test checks custom M2540A hardware. The basic test executes in less than five
You can close the RST application using either of the following two methods:
When the RST application closes, a window asks if you want to reboot.
minutes and requires no operator interaction while executing.
➤Closing the RST Application
➤Running a Basic Test

To run a basic test:

The extended tests check the system’s hardware to the board level, and do not require your
participation after they are started.
➤Running an Extended Test
To run an extended test:


Figure 6-10 RST Extended Tests

using the tree control in the left side of the RST window. As an example, Figure 6-11 shows the
Each subsystem test consists of test modules, and each test module has submodules. You can
select individual modules or submodules to run separate from the other tests in its group by
BP/AP subsystem test’s Demodulator test module and the module’s submodules.

Figure 6-11 Demodulator Tests (Example)

display, in its Category column, and in the Status window at the lower left of the window. See
The two-digit codes that precede the submodule names are used in the Application Test Log
The interactive tests check the operation of the system’s controls. They require you to make
selections and operate controls as they execute.
➤Running an Interactive Test
“RST Test Log” on page 121.
To run an interactive test:

menu.

The logs list errors in chronological order, with the most recent errors at the top of the list. You
The RST application logs errors during operation of the system and when you run a test. View
the logs by clicking Event Viewer under the Log tab of the RST window and selecting the log
This log includes run-time errors encountered by both the ultrasound application and the RST
application. The log lists the date and time the error occurred, the name of the error, the
can view more detail of an error by double-clicking the entry in the list view.

The Event Viewer includes application logs, a security log, and a system log.
you want to see. Figure 6-12 shows the RST log display.
➤Viewing Error Logs
Figure 6-12 RST Log Display

applications and subsystems affected, and other information. Figure 6-13 shows the ultrasound

To clear the Ultrasound Application Log, complete the following procedure:
➤Clearing the Ultrasound Application Log
Figure 6-13 Ultrasound Application Log Display
application log display.

This log includes parameters of past RST tests and errors encountered while running the tests.

Figure 6-14 shows the Event Viewer application RST test log.
Figure 6-14 RST Application Test Log Display

To find the meaning of the numeric codes, use the tree control in the RST window’s left side as
The example indicates that the A/D converter (submodule 02 of module 01) of the display
The Application Test Log’s Category column contains coded information. The code is in the
described in “Test Modules” on page 117.
➤Clearing the RST Test Log
processor was tested.
following format:
Where:
SSnnNN

To view detailed test results, open the Event Properties window as follows:

Figure 6-15 Event Properties Window (Example)
The Event Properties window opens.
➤Viewing Test Details

The Event Viewer system log includes hardware errors filtered from the application run-time
error log and the RST test log. Figure 6-16 shows the Event Viewer system log.

➤Clearing the System Log
Figure 6-16 System Log Display

You can enter information in the service log that is displayed under the Log tab. You must log on
to the RST application using a valid service name and password to make an entry in the service
Service Log The service log on the Log tab’s tree is a list of comments entered by service personnel.
➤Making an Entry in the Service Log
To make a new entry in the service log:


log.

You can change the information in an existing service log entry. You must log on to the RST
application using a valid service name and password to modify an entry in the service log.

➤Modifying an Entry in the Service Log
To modify a service log entry:

available. Submenus of the OSD Main Menu control the size and geometry of the display and set
The four buttons on the front of the monitor adjust brightness and contrast. When the two
right-hand buttons are pressed together as the monitor powers up, the OSD Main menu is
the monitor’s signal input voltage (see “Monitor Signal Voltage” on page 62).
Pressing the two right-hand buttons together degausses the CRT.
OSD Main Menu ➤Opening the OSD Main Menu
The OSD Main Menu displays.
Monitor

Vertical Size For viewing BIOS settings mode, you must adjust the VGA vertical size. This adjustment does
The four buttons on the front of the monitor control all the OSD functions, as follows:
➤Adjust the VGA Vertical Size
value of the selected parameter.
not affect the normal display mode.
The OSD Main menu displays.
parameter.

The vertical size adjustment bar displays.
The vertical size adjustment bar closes.
The Size & Position menu displays.
The SIze & Position menu closes.


Boot Sequence ➤Change the Boot Sequence
The Device menu opens.

The air filter resides inside the service panel at the right front of the lower enclosure, in front of
the fan. If the filter becomes very dirty, it can impede airflow, which may cause overheating of the
Preventive Maintenance:
During preventive maintenance, look for conditions that may cause problems, and loose knobs
This section contains procedures for cleaning the M2540A and its components.
cleaners may cause the colors to fade. Use only mild soap.
To remove and clean the air filter:
➤Cleaning the Air Filter
power supply or the e-box.
Safeguards” on page 139.
or worn TGC controls.

Preventive Maintenance:
Inspect the interior of the e-box for dust (See “Replacing E-box Circuit Boards” on page 151),
➤Cleaning the Trackball
and vacuum as necessary.
unscrewing the bezel.
➤Cleaning the e-box
To clean the trackball:
material.

Preventive Maintenance:
protected surface. Use adequate ESD safeguards when performing any service or maintenance
removing them from the system. After removal, make sure the boards lie flat on an ESD
To degauss the monitor, press the two right-hand monitor buttons at the same time.
Perform all RST and system checks to verify performance.
vacuuming. See “ESD Safeguards” on page 139.
procedure. See “ESD Safeguards” on page 139.
➤Degaussing the Monitor

This section contains tips and procedures for finding the causes of malfunctions of the M2540A.
power supply is energized by line voltage from the mains. The LED takes a few seconds to
power supply by jumping the PWR DOWN pin on the BPAP board to chassis ground.
If the M2540A does not start when the PC’s power button is pressed, proceed as follows:
Always check that power cords are plugged in, and inspect them for wear.
For problems with optional peripherals, see the Peripherals Manual.
is faulty. See “PC Service” on page 142.
cord to connect the PC to mains.
light when the system is first plugged in.
Boot

If the M2540A’s internal PC starts, and the monitor functions normally, but one or more
If the M2540A’s internal PC starts, but the monitor stays black, proceed as follows:
peripheral devices don’t work, proceed as follows:
power up, it is faulty. Replace the device.
page 154.
page 157.
page 154.
PC Boots, but
Peripherals
Don’t Work
Monitor is
Blank

lation Problems If a connected printer does not show up on the Setup/Peripherals screen, perform the following
HP940 Startup The first time the HP940 printer is powered up, it may cause the system to hunt for more
If the procedure does not solve the problem, contact your service representative.
peripherals to install. Rebooting the system stops this activity.
indicators.
procedure:
Problems
Printer

Electromagnetic interference (EMI) can degrade the performance of the M2540A. Answering the
A transducer placed close to an ECG cable can increase interference. Moving the ECG cable
Answers to these questions help determine whether the problem is with the system or with the
scanning environment. Contact your local Philips service representative with a full description
or other medical equipment away from the transducer can reduce interference.
following questions can help locate the source of interference:
and your observations.
Interference
netic

Overview This chapter includes all the service procedures that apply to the M2540A. To locate parts and
assemblies referred in the procedures, see the exploded diagrams and parts lists in “Parts” on
This manual does not include service procedures for optional peripherals. See the Philips
Ultrasound Systems Peripherals Guide for that information.
The chapter includes these sections:
page 177.

Follow the electrical static discharge (ESD) guidelines in this section to prevent deterioration or
Most components in the M2540A are replaced as modules. The power supply, the monitor, and
Some parts are available at reduced cost, as exchange items. See Table 13-12 on page 200 for a
the system control panel are not serviceable, and are replaced as units. The circuit boards and
surfaces. A protective membrane under the keyboard provides some protection against spills,
electricity, take the necessary ESD precautions to avoid static buildup and discharge.
Liquids, including ultrasound gel, can adversely affect the system control panel and other
but liquids may seep into electrical components under the panel and cause problems.
function cards inside the M2540A’s internal PC and e-box are replaceable in the field.
To prevent damage to the system, keep the following good practices in mind:
destruction of sensitive electronic components. General guidelines include:
list of exchange parts and their part numbers.
Philosophy
Safeguards
General
Repair
ESD

You can service components mounted above the lower enclosure without opening any panels in
To service components mounted in the lower enclosure, you must open one or more panels on
This section describes how to gain access to the serviceable components of the M2540A system.
the lower enclosure. For locations of the panels and components, see Figure 13-3 on page 185.
the lower enclosure. These components include the following:
The components in the lower enclosure include the following:
Components in
Components
Above the
the Lower
Servicing
Servicing
Enclosure
Service
Access

either panel, remove the three screws along its lower edge, pull the bottom of the panel out, and
You can make notes in the RST application’s service log, to help with later service procedures. To
Doors Two doors in the rear of the lower enclosure give access to the rear of the PC, the e-box, and
Panel A panel at the right side of the system’s front gives access to the transducer connector module
Side Panels The two lower enclosure side panels are attached with screws on their lower edge. To remove
read service log entries, see “Using the Service Log” on page 125. To make an entry in the RST
the power supply. The doors are hinged near the center of the system, and are held closed by
open this panel, press down the two latches at its top, pull the top edge out, and lift the panel
mounting, the air filter, the fan, the front of the e-box, and the front of the power supply. To
quarter-turn latches near the sides of the system. To open the doors, turn the latches
service log, see “Making an Entry in the Service Log” on page 125.
lower the panel clear of the system.
counter-clockwise.
clear of the system.
system.
Rear Service
Record
Keeping

PC Service This section contains procedures for servicing M2540A software and the hardware components
The following procedures describe reloading and upgrading the ultrasound software, and backing
Complete the following procedure to reload or upgrade the system software. The procedure
the MOD drive are empty before beginning software installation.
takes about 45 minutes to complete and requires three CDs:
➤Reloading or Upgrading System Software
up and restoring customer presets.
of the internal PC.
Software
Service

installation. Do not press a key when it does. The message clears after a few moments.
Before performing any service that deletes a customer’s presets, back them up to a floppy disk.
network settings. The backup function records settings made on the DICOM Setup
window, but does not record those settings made on the Network Settings window.
When the COTS software installation finishes, the system automatically restarts. If the
After the procedure is complete, restore the customer’s presets from the floppy disk.
installation is successful, the monitor displays the ultrasound imaging screen.
➤Backing Up Customer Presets
To back up a customer’s presets:
Press Y.



system settings, the peripheral settings, and the installed options.



➤Restoring Customer Presets
To restore a customer’s presets:

require replacement. You can replace some PC components without removing the PC from the
replace other components, you must remove the PC from the system. (See “Removing the PC
matches the serial number of the system to that of the system that created the disk.
system, by opening the side of the PC. (See “Opening the PC Side Cover” on page 148.) To
The PC contains the following replaceable components (shown in Figure 13-1, on page 180):
PC Components None of the PC’s internal components are adjustable or repairable. Faulty PC components
from the System” on page 150.)

b. Gently pull the four plastic latches at the left side of the PC’s bezel, while pulling the left
You can gain access to components in the PC without removing the PC from the system, by
To open the PC side cover, see Figure 13-3 on page 185 to locate parts and assemblies, and
a. Slide the PC two cm to the left, away from the center of the system.
side of the bezel forward until the latches clear the PC case.
wrist band, and work on a static-dissipating mat if possible.
➤Opening the PC Side Cover
complete the following procedure:
opening the PC side cover.

d. Each of the replaceable parts listed in step 6 is retained by two plastic latches. To remove
the part, press the two latches in toward the part, and slide the part out of the PC case.
Reverse the procedure to reassemble the PC. Perform all RST and system checks to verify
c. Slide the bezel to the right until it can be removed.
performance.

To remove the PC, see Figure 13-3 on page 185 to locate parts and assemblies, and complete the
To install the PC in the system, reverse the procedure above. Perform all RST and system checks
➤Removing the PC from the System
to verify performance.
following procedure:

E-box You can perform most e-box service without removing the e-box from the system. To perform
grounded wrist band, and work on a static-dissipating mat if possible. See “ESD Safeguards” on
service that does require removing the e-box, you must also remove the power supply (See
To replace a circuit board in the e-box, see Figure 13-2 on page 183 to locate parts and
“Removing the Power Supply and the E-box” on page 154).
assemblies, and complete the following procedure:
➤Replacing E-box Circuit Boards
disconnecting the cable.
page 139.

To replace the e-box motherboard, See Figure 13-2 on page 183 to locate parts and assemblies,
grounded wrist band, and work on a static-dissipating mat when possible. See “ESD Safeguards”
Power Supply and the E-box” on page 154 and “Separating the E-box and the Power
with the e-box removed and separated from the power supply. See “Removing the
and complete the following procedure:
➤Replacing the Motherboard
Supply” on page 155.
on page 139.

grounded wrist band, and work on a static-dissipating mat if possible. See “ESD Safeguards” on
“Removing the Power Supply and the E-box” on page 154 and “Separating the E-box
To replace the distribution board, see Figure 13-2on page 183 to locate parts and assemblies,
are easier with the e-box removed and separated from the power supply. See
and the Power Supply” on page 155.
➤Replacing the Distribution Board
and complete the following procedure:
to the distribution board’s connectors.
page 139.

Power Supply The M2540A power supply is not serviceable in the field, except for replacing the fan. The power
supply and the e-box are bolted together and must be removed from the system before they can
cable is plugged into a power source. If the PC is connected to one of the other outlets, it
To remove the power supply and the e-box from the system, see Figure 13-3 on page 185 to
be separated. Replacement power supplies are supplied with the fan and the finger guard
locate parts and assemblies, and complete the following procedure:
➤Removing the Power Supply and the E-box
does not boot up.
installed.

grounded wrist band, and work on a static-dissipating mat if possible. See “ESD Safeguards” on
➤Separating the E-box and the Power Supply
To separate the e-box from the power supply:
page 139.

To remove the fan, see Figure 13-2 on page 183 to locate parts and assemblies, and complete the
The screws that secure the transducer connector modules are behind the service panel on the
Fan The M2540A’s cooling fan is attached to the front of the power supply. Replacement fans are
To replace a transducer connector module, see Figure 13-2 on page 183 to locate parts and
➤Replacing Transducer Connector Modules
assemblies, and complete the following procedure:
to the distribution board’s connectors.
supplied with the finger guard installed.
front of the lower enclosure.
➤Replacing the Fan
following procedure:
Connector
Modules

field, so a monitor that does not perform properly must be replaced. See “Monitor Adjustment”
Monitor The color monitor is a critical component of the system. The monitor is not serviceable in the
To remove the monitor from the system, see Figure 13-5 on page 193 to locate parts and
on page 127 for information on adjusting the monitor display.
rivet bodies with pliers may ease installation.
assemblies, and complete the following procedure:
➤Replacing the Monitor

The M2540A twivel is not adjustable. The following procedure describes how to disassemble the
To reassemble the twivel assembly, perform the above procedure in reverse. When completing
step 3, pass the VGA cable’s connector through the twivel base first, before passing the power
Except for the trackball (see “Cleaning the Trackball” on page 132), the system control panel
To disassemble the twivel, see Figure 13-5 on page 193 to locate parts and assemblies, and
Perform all RST and system checks to verify performance.
base. Note which hole the cables go through.
contains no serviceable parts, and is replaced as a unit.
complete the following procedure:
➤Disassembling the Twivel
twivel to replace its parts.
before the video cable.
cord connector through.
Monitor
System
Twivel

To replace the system control panel, see Figure 13-5 on page 193 to locate parts and assemblies,
the control panel standoffs. The standoffs will cut any cables underneath them when the screws
band, and work on a static-dissipating mat if possible.
➤Replacing the System Control Panel
and complete the following procedure:
are tightened.

The angular enclosure on top of the control panel surround is the monitor support. To replace
To replace the monitor support and its front panel, see Figure 13-5 on page 193 to locate parts
the mechlok cable, lift column slides, upper cable harness, or lower enclosure top panel, you
Cables Most of the cables in the M2540A system pass through the lower enclosure’s central plenum.
You may have to remove the panels from the rear of the plenum when replacing a cable.
which hole the cables are in as you take the support off.
➤Replacing the Monitor Support and its Front Panel
and assemblies, and complete the following procedure:
must remove the monitor support and its front panel.
Front Panel
Monitor

bundled together. To replace any one of these cables, you must remove the entire bundle. The
Cables that connect the lower components to the system control panel and the monitor are
➤Replacing the Upper Assembly Cable Bundle
following cables comprise the bundle:
To replace the cable bundle:
Monitor Cables
Panel and
system control panel · control panel · control panel


Although the front wheel assemblies on the cart are different from the rear wheels (front wheels
have swivel locks and brakes, rear wheels do not lock), the replacement procedure for all wheels
Each cart wheel is supplied as a caster assembly. The front caster assembly includes the integral
wheels at that end slightly off the floor. Make sure the cart is stable in this raised position.
Cart This section contains procedures for servicing the lift column, the wheels, and the lower
➤Replacing Cart Wheels (Caster Assemblies)
rolling during this procedure.
To replace a caster assembly:
brake and swivel lock.
enclosure top panel.
is the same.

To replace the top panel of the lower enclosure, see Figure 13-3 on page 185 to locate parts and
You can remove and install the top panel of the lower enclosure without taking out the lift
➤Replacing the Top Panel of the Lower Enclosure
assemblies, and complete the following procedure:
column or the upper cable bundle.
cable ends’ positions.

service procedures. Because the lift column supports the weight of the upper system, you must
To prepare the system for lift column servicing, see Figure 13-4 on page 190 and Figure 13-3 on
The column is also available as an assembly, including all the components listed above. After
reassembling the system, perform all RST and system checks to assure proper performance.
The procedures for replacing the lift column components are the most complex of the field
complete the following procedure before performing any service to the column.
page 185 to locate parts and assemblies, and complete the following procedure:
➤Preparing the System for Servicing the Lift Column
Lift Column The lift column has four replaceable components:
the lift column.

The gas spring supports the weight of the upper assembly when the system is upright. Before
working on the gas spring, you must relieve the weight.
➤Replacing the Gas Spring
and damage to the system.
To replace the gas spring:

The mechlok assembly locks the lift column at any height in the column’s range. The mechlok is
surround. You can replace the cable without completely removing the mechlok, and you can
controlled by a cable connected to a handle under the front of the system control panel
replace the mechlok mechanism without removing the cable from the system.
➤Replacing the Mechlok and Cable
To remove the mechlok and cable:
and damage to the system.


When replacing slides, it is not necessary to remove the mechlok cable, or to disconnect the
Each of the three lift column slide assemblies consists of an inner slide mounted on the lift
column and an outer slide mounted in the lower enclosure plenum.
mechlok or the gas spring from their lower bracket.
To replace the lift column slides:
➤Replacing the Slides
and damage to the system.

ball carriers.

Cabling:
The following table lists the cables in the M2540A. The cables may also be listed in assembly
Description Part Number Qty Reference
Cart to external AC power cable M2540-69120 1 PC (DP) power cable 8120-2572 1 Table 13-5
Internal LAN cable M2540-69130 1 Table 13-5
Composite video RCA cable M2540-69140 1 Trigger miniphone cable M2540-69150 1 Table 13-5
Printer USB cable M2540-69100 2 Peripheral power cable (IEC) 8120-2572 3 Table 13-5
I/O panel foot switch cable M2540-69160 1 Table 13-5
Token ring cable M2540-69000 1 Table 13-5
Ground cable M2540-69170 1 System cable harness M2540-69220 1 Table 13-5
Monitor power cable M2540-69180 1 Monitor video cable M2540-69040 1 Speaker cable M2540-69010 1 Keyboard USB cable M2540-69110 1 Keyboard auxiliary power cable M2540-69050 1
parts lists shown in the “Reference” column of this table.
Table 11-1 Cables

Cabling:
Description Part Number Qty Reference
Keyboard line audio cable M2540-69060 1 Power cords
UK power cord M2540-60900 1 Australia power cord M2540-60901 1 Europe power cord M2540-60902 1 US, Canada, and Japan power cord M2540-60903 1 Switzerland power cord M2540-60906 1 Denmark power cord M2540-60912 1 South Africa power cord M2540-60917 1 Israel power cord M2540-60919 1 Argentina power cord M2540-60920 1 PRC power cord M2540-60922 1 Physio option cables
Physio option USB cable M2540-69020 1 Physio option trigger cable M2540-69030 1 3-Lead AAMI grabber set M1603A
3-Lead AAMI trunk cable M1500A
3 Lead AAMI mini-clip set M1609A
3-lead IEC trunk cable M1510A
Table 11-1 Cables (Continued)

Cabling:
Description Part Number Qty Reference
VCR option video in/out cable M2540-69070 2 VCR option audio in/out cable M2540-69200 2 VCR option serial cable M2540-69090 1 Optional accessory cables
External LAN cable M2540-69080
USB cable 10-foot M2540-69110
Table 11-1 Cables (Continued)
VCR option cables

information on the correct settings. The hard disk drive is the only device on the primary IDE,
The jumper positions and switch settings of the drives in the M2540A depend on the
Jumpers Figure 12-1shows typical hard disk drive jumper positions. Examine the drive itself for

and is usually set to MASTER or SINGLE.
Figure 12-1 Hard Disk Drive Jumper Settings
manufacturer of the drives.
Settings

Jumpers Figure 12-2 shows typical CD-RW drive jumper positions. Set the jumper to the MASTER
Jumpers Set the MO drive jumpers to SLAVE on the secondary IDE.
The VCR’s switches must be set to the following positions:
Figure 12-2 CD-RW Drive Jumper Position
position on the secondary IDE.

CD Drive
Settings

On the Sony black-and-white printer, dip switch number 5 should be up. All the other dip
switches should be down.
Sony B/W
Switch
Settings

Overview This section contains lists of the M2540A’s parts, supplies, and accessories available for use with
Address all parts orders or inquiries to your local Philips Sales and Service office. Always specify
the correct part number or exchange number for the assembly or component being ordered, as
well as the serial and revision numbers of the system in which the assembly is being installed.
the system, and information on ordering parts, accessories, and supplies.
Parts

New Parts Be sure to specify the correct sub-assembly or component part number, as listed throughout this
Exchange Parts Some circuit boards, transducers, and OEM peripherals are available as exchange items (repaired
rear of the cart. The ID label also identifies installed system options. Use the system Help file to
or refurbished), at reduced cost. Exchange item part numbers are different from original equip-
When ordering any new or exchange part, be sure to include the system serial number and the
software revision prefix (for example, B.0). The serial number is on the system ID label, on the
The following table lists the major assemblies of the M2540A, and the drawings that show them.
find software revision numbers, or run the RST application, and under the Info tab, click Sys-
To learn about transducer parts, please see the M2540A Transducer Guide.
PC Figure 13-1 on page 180
Power supply Figure 13-2 on page 183
e-box Figure 13-2 on page 183
Upper assembly Figure 13-5 on page 193
Lower assembly Figure 13-3 on page 185
Lift column Figure 13-4 on page 190
ment part numbers. See “Exchange Part Numbers” on page 200
Description Drawing and Page
tem. (See “System Information” on page 109.)
Table 13-1 Major Assemblies
section.
Revision Level
Number and
Assemblies
Major

The following table lists the available transducer connector modules and their part numbers.
Description Part Number Qty
Cannon transducer connector module M2540-62100 1 Advance Vision transducer connector module M2540-62000 1 Pencil transducer connector module M2540-62200 1 Blank module M2540-62010 1
Table 13-2 Transducer Connector Modules
Connector
Modules

Internal PC The following figure shows the replaceable components of the M2540A’s PC. Numbers refer to

Figure 13-1 Internal PC Components
Table 13-3, on page 181.

The following table lists the replaceable components of the internal PC shown in Figure 13-1.
Numbers in the # column are the item numbers on the drawing. A dash (—) in the # column
# Description Part Number Qty Reference
PC M2540-67000 1 Figure 13-3
1 APIO board M2540-60030 1 2 PC motherboard M2540-67110 1 3 CPU M2540-67140 1 4 Hard disk drive M2540-67100 1 5 CD-RW drive M2540-67130 1 6 Floppy disk drive M2540-67120 1 7 PC power supply M2540-67150 1 8 PA 810 enclosure M2540-67170 1 9 PC bezel M2540-67210 1 — Ribbon Cable kit M2540-67160 1 — 256 MB memory module M2540-67180 1 — 512 MB memory module M2540-67190 1 — Drive bay blank cover kit M2540-67200 1 — Screw, 6-32 x 5/8, panhead 2360-0123 3
Table 13-3 Internal PC Components
means the part is not numbered on the drawing.

# Description Part Number Qty Reference
Optional physio module Table 13-11
10 Physio module M2540-62300 1 11 Physio USB cable M2540-69020 1 Table 11-1
12 Physio trigger cable M2540-69030 1 Table 11-1
13 MOD drive M2540-80100 1 — MOD adapter assembly M2540-80400 1 PC components for VCR option — — Table 13-9
14 Video converter board M2540-80070 1 Table 13-9
15 Video capture board M2540-80080 1 Table 13-9
16 Sound card M3150-60020 1 Table 13-9
Table 13-3 Internal PC Components (Continued)
Optional magneto optical drive — —

The following figure shows the system power supply and the components of the e-box listed in

Table 13-4. The power supply contains no replaceable parts.
Figure 13-2 e-box and Power Supply Components
e-box and

Figure 13-2. Numbers in the # column are the item numbers on the drawing. A dash (—) in the
The following table lists the components of the power supply and e-box assembly shown in
2 Cooling fan M2540-66010 1 3 e-box assembly M2540-64000 1 Does not include circuit boards
1 Power supply M2540-66000 1 Exchange part, see Table 13-12
4 Motherboard M2540-60000 1 5 Distribution board M2540-60010 1 6 TR board M2540-60040 2 Exchange part, see Table 13-12
7 Demodulator board M2540-60050 1 Exchange part, see Table 13-12
8 BPAP board M2540-60060 1 Exchange part, see Table 13-12
Table 13-4 Power Supply and e-box Assembly Components
# Description Part Number Qty Reference
# column means the part is not numbered on the drawing.
9 Chassis septum M2540-00040 2 10 Card guide 4040-2551 8 11 RFI shielding kit M2540-86040 1 12 M3 x 6mm screw 0515-0430 27 13 M4 x 10-mm screw M2540-80322 16 14 e-box door assembly M2540-64010 1

The following figure shows the components of the lower assembly listed in Table 13-5.

Figure 13-3 Lower Assembly
Lower

The following table lists the components of the lower assembly shown in Figure 13-3. Numbers
in the # column are the item numbers on the drawing. A dash (—) in the # column means the
2 Air filter M2540-80050 1 3 e-box and power supply — 1 Bolted together, see Figure 13-2
8 Side panel, right M2540-00400 1 9 Side panel, left assembly M2540-65190 1 10 Rear panel assembly M2540-65170 1 Includes rear doors
# Description Part Number Qty Reference
5 Control top — 1 Figure 13-5
6 VCR — 1 Table 13-13
7 Printer — 1 Table 13-13
Cart assembly M2540-65000 1 1 PC M2540-67000 1 Table 13-3
4 Column assembly M2540-65100 1 Table 13-6
11 Cart top M2540-00420 1 12 Center column, front M2540-00370 1 13 Latch, rear door 1390-1326 2 14 Front panel M2540-00440 1 15 Top front trim M2540-00590 1 16 Upper I/O panel M2540-00660 1
Table 13-5 Lower Assembly
part is not numbered on the drawing.

# Description Part Number Qty Reference
17 Lower I/O panel M2540-00670 1 LAN bulkhead connector 1252-6178 1 Grounding stud 1251-5964 1 I/O panel trigger cable M2540-69150 1 Table 11-1
I/O panel foot switch cable M2540-69160 1 Table 11-1
Internal LAN cable M2540-69130 1 Table 11-1
Composite video cable M2540-69140 1 Table 11-1
18 Cart front panel bracket M2540-00620 1 19 Power supply bracket M2540-00530 2 20 PC bracket M2540-00600 1 21 Transducer cable hook M2540-40690 4 22 Cable hook, rear M2540-40680 1 23 External power plug and cable M2540-69120 1 24 Peripheral strap 77922-80290 1 25 Strap bracket kit M2540-86020 1 — Color printer tray kit M2540-86010 1 26 VCR tray M2540-00810 1 — Printer spacer M2540-40750 4 — Nylon bushing, 2-inch 0040-0951 1 — VCR option video in/out cable M2540-69070 2 Table 13-9
— VCR option audio in/out cable M2409-69200 2 Table 13-9
Table 13-5 Lower Assembly (Continued)

# Description Part Number Qty Reference
— VCR option serial cable M2540-69090 1 Table 13-9
— Printer USB cable M2540-69100 2 — Peripheral power cable (IEC) 8120-2572 3 Table 11-1
Table 13-5 Lower Assembly (Continued)
— M3 x 12-mm screw 0515-0664 10 28 M4 nylock nut 0535-1029 12 — Foam strip M2540-80120 4 29 Center cart assembly M2540-65140 1 30 Plain swivel caster M2540-80240 2 31 Locking swivel caster M2540-80250 2 32 M8 x 16-mm hex locking bolt 0515-2466 16 — M4 x 10-mm screw with
— Token ring cable M2540-69000 1 — Ground cable M2540-69170 1 — Power cord M2540-40700 1 27 M4 x 10-mm Phillips locking
— M6 lock nut 0535-1030 4 — M5 x 16-mm screw 0515-0671 1 — M5 plastic hex nut 0535-0090 1 — Nameplate BASIC M2540-83010 1
washer M2540-80322 39 · screw 0515-2859 4

# Description Part Number Qty Reference
Table 13-5 Lower Assembly (Continued)
— Nameplate Performance M2540-83020 1 — System Serial Overlay Label M2540-83030
— Performance label strip M2540-83040 1 — Nameplate CV M2540-83050 1 — Nameplate CV performance M2540-83060 1

The following figure shows the components of the lift column assembly listed in Table 13-6.

Figure 13-4 Lift Column
Assembly
9 15 12

The following table lists the components of the lift column assembly shown in Figure 13-4. Num-
bers in the # column are the item numbers on the drawing. A dash (—) in the # column means
6, 7, and strut eyelet 4 Mechlok bracket — 1 Part of
1 Column assembly M2540-65100 1 2 Slide assembly M2540-65030 3 3 Gas spring bracket M2540-00310 1 — Mechlok bracket hardware kit M2540-65180 — Includes items 4, 5,
M2540-65180
M2540-65180
M2540-65180
# Description Part Number Qty Reference
5 Grommet, 5/16-inch ID — 4 Part of
6 Flat washer, 5/16-inch — 4 Part of
7 Clevis pin — 2 Part of
8 Gas strut assembly M2540-80020 1 9 Column cover M2540-00430 1 10 Mechlok M2540-80030 1 11 Slide bracket, front M2540-00320 1 — Mechlock cable M2540-80035 1
Table 13-6 Lift Column Components
the part is not numbered on the drawing.

# Description Part Number Qty Reference
Table 13-6 Lift Column Components (Continued)
— Friction ring M2540-80010 2 — Belleville washer 3050-1403 8 12 Detent ring M2540-00270 1 13 Cable protector cap M2540-40710 1 14 Quick release pin and ring 1480-1141 1 15 Pivot mount M2540-00230 1 16 M4 x 6-mm panhead screw 0515-0684 12 — M4 x 10-mm screw with washer M2540-80322 12 — M4 x 10-mm screw with washer,
— M4 x 10-mm screw with washer, light gray M2540-80325 2 17 M4 lock nut 0535-1029 12
mushroom M2540-80320 10

The following figure shows the components of the upper assembly listed in Table 13-7.

Figure 13-5 Upper Assembly
Upper

column are the item numbers on the drawing. A dash (—) in the # column means the part is not
Table 13-7 lists the components of the upper assembly shown in Figure 13-5. Numbers in the #
1 Top shelf assembly 1 Figure 13-6 on page 196
# Description Part Number Qty Reference
6 Cable stop bracket M2540-00550 1 7 M4 x 16-mm flathead screw 0515-4620 12 8 M4 x 25-mm flathead screw 0515-4622 25 9 M4 x 25-mm panhead screw 0515-2860 2 10 Transducer cable hook M2540-40690 3 — Twivel assembly
— FIMI monitor, mushroom M2540-66500 1 2 System control panel (keyboard) M2540-40100 1 3 Height release handle M2540-40560 1 4 Monitor support M2540-00210 1 5 Monitor support front panel
11 Twivel ring M2540-40740 1 12 Twivel shoe M2540-40730 1 13 Monitor mount (twivel base) M2540-40630 1 14 Twivel hat M2540-40610 1
Table 13-7 Upper Assembly Components
numbered on the drawing.
assembly M2540-65120 1

22 M4 nylock nut 0535-1029 2 — M4 x 10-mm screw with washer M2540-80322 2 23 Trackball assembly M2540-80200 1 Includes items 24 and 25
— Pivot mount assembly 1 — Pivot mount M2540-00230 1 15 Pivot knob M2540-40550 1 16 Latch cam M2540-40590 1 17 Pivot latch assembly M2540-65150 1 18 Latch bearing M2540-40600 1 19 M4 x 6-mm Phillips screw 0515-2859 3 Used with items 20, 21
# Description Part Number Qty Reference
Table 13-7 Upper Assembly Components (Continued)
24 Trackball M2540-80210 1 25 Trackball ring M2540-80220 1 — Control panel replacement knob kit M2540-86070 —
— Rotary encoder M2540-40210 7 — Cable tie 1400-0401 12
20 Delrin flat washer 3050-2184 3 21 Precision flat washer
0.171-inch ID 3050-2186 3


Top Shelf The following figure shows the components of the upper assembly listed in Table 13-8.
Figure 13-6 Top Shelf

The following table lists the parts shown in Figure 13-6. Numbers in the # column are the item
numbers on the drawing. A dash (—) in the # column means the part is not numbered on the
1 Bottom sheet M2540-00160 1 2 Keyboard surround M2540-40500 1 3 Top sheet M2540-00150 1 4 Speaker support, left M2540-00190 1 5 Speaker support, right M2540-00200 1 6 Front panel retainer clip M2540-00880 4 7 Speaker M2540-80230 2 8 P-clip, 1-inch 2 Part of M2540-86050
9 P-clip, 5/8-inch 1 Part of M2540-86050
10 P-clip, 3/16-inch 2 Part of M2540-86050
# Description Part Number Qty Reference
11 M4 x 10-mm Phillips screw 0515-2859 4 12 M3 x 6-mm Phillips screw 0515-0430 8 — Speaker foam M2540-80110 2
Table 13-8 Top Shelf Components
drawing.

The following table lists the parts of the VCR option. For part numbers of available VCRs, see
Software The following table lists the part numbers for the M2540A’s software CDs.
Description Part Number Qty
Ultrasound application CD M2540-81000-01
Operating system CD M2540-82500-01
Video converter board M2540-80070 1 Video capture board M2540-80080 1 Sound card M3150-60020 1 Video in/out cable M2540-69070 2 Audio in/out cable M2450-69200 1 Serial cable M2540-69090 1
Table 13-10 M2540A Software CDs
Table 13-9 VCR Option Parts
“Optional Peripherals” on page 201.
Parts
Description Part Number

Description Part Number Qty Reference
Physio module (ECG module) M2540-62300 1 Table 13-3
Physio USB cable M2540-69020 1 Table 11-1
Physio trigger cable M2540-69030 1 Table 11-1
ECG electrodes 77020-62180 Pediatric electrodes, disposable 77020-62190
The following table lists the parts of the physio option.
3-Lead AAMI grabber set M1603A
3-Lead AAMI trunk cable M1500A
3 Lead AAMI mini-clip set M1609A
3-lead IEC trunk cable M1510A
3-lead IEC grabber set M1613A
3-lead IEC mini-clip set M1619A
Table 13-11 Physio Option Parts
Parts

Some parts are available at reduced cost, as exchange items. The following table lists the part
Description Part Number Exchange Part
FIMI monitor M2540-66500 M2540-66508
System power supply M2540-66000 M2540-66008
TR board M2540-60040 M2540-60048
Demodulator board M2540-60050 M2540-60058
BPAP board M2540-60060 M2540-60068
NTSC VCR 77616A 77616-68000 PAL VCR 77626A 77626-68000 Black-and-white printer UPD-895 (USB) M2540-66700 M2540-66708
Color printer UPD-21MD (USB) M2540-66800 M2540-66808
S4 ultraband sector transducer 21330A 21330-68000 S8 ultraband sector transducer 21350A 21350-68000 S12 transducer 21380A 21380-68000 c3540 biopsy capable transducer 21321A 21320-68000 EC6509 endocavity transducer 21336A 21336-68000 L7535 linear transducer 21359A 21359-68000
Table 13-12 Exchange Part Numbers
numbers for exchange parts and assemblies.
Numbers
Keyboard M2540-40100 M2540-40108 · Number

NTSC VCR 77616A Exchange part, see Table 13-12
PAL VCR 77626A Exchange part, see Table 13-12
Black-and-white printer UPD-895 (USB) M2540-66700 Exchange part, see Table 13-12
Color printer UPD-21MD (USB) M2540-66800 Exchange part, see Table 13-12
Description Part Number Exchange Part
The following table lists the part numbers of peripheral options available for the M2540A.
L5035 linear transducer 21360A 21360-68000 TEE (Omniplane II) transducer 21369-60000 21369-68000 L1038A small parts transducer 21376A 21376-68000 15-6L intraoperative transducer 21390A 21390-68000 PA 4-2 sector transducer 21422A 21422-68000 CA 5-2 transducer 21425A 21425-68000
Description Part Number Note
Table 13-12 Exchange Part Numbers
Color plain-paper printer HP 940C
Black-and-white plain paper laser printer HP1200
Table 13-13 Optional Peripherals
Optional
Number · MOD drive M2540-80100

The following table lists supplies and accessories available from Philips for use with the M2540A.
To order supplies and accessories, see “Ordering Parts” on page 177. To learn about transducer
Microphone 9164-0719 Foot switch M2540-80280
External LAN cable M2540-69080
USB cable 10-foot M2540-69110
Table 13-14 Supplies and Accessories
accessories, please see the M2540A Transducer Guide.
compliance.
Accessories

black-and-white Sony digital graphic printer UP-D895 SNY110HG
black-and-white Sony digital graphic printer UP-D895MD SNY110HD
Sony UPP-110S standard paper 5 roll box SNY110S
Aquasonic gel, US, 12 8.5-oz. bottles 40483A Aquasonic gel, US, 1 5-liter-bottle refill 40483B Ultrasound gel, Europe, 12 bottles 40404A Ultrasound gel, Europe, 1 5-liter-bottle refill 40404B
Table 13-14 Supplies and Accessories (Continued)
Sony UPC 21 L color printing pack for the Sony digital color
Sony UPP-110 HD high density paper printing pack for the
Sony UPP-110 HG glossy paper printing pack for the

Biopsy kit C3540 Civco 645-057
Biopsy kit PA4-2 Civco 645-069
Biopsy kit CA5-2 Civco 645-067
Biopsy kit L1038 Civco 645-059
Biopsy kits L7535 and L5035 Civco 645-039
Biopsy kit E6509 Civco 610-588
Foam electrodes, pregelled, 100 packs 40420A Pediatric electrodes, disposable 13951C ECG Lead Sets
3-lead set, grabber, AAMI M1603A
3 lead trunk cable, AAMI M1500A
3-lead set, snap, AAMI M1605A
3-lead set, mini-clip, AAMI M1609A
3 lead trunk cable, IEC M1510A
3-lead set, grabber, IEC M1613A
3-lead set, snap, IEC M1615A
3-lead set, mini-clip, IEC M1619A
Table 13-14 Supplies and Accessories (Continued)
Description Part Number · ECG electrodes · Biopsy kits

Transducers:
exchange items. See Table 13-12 on page 200 for a list of exchange part numbers.
PA 4-2 sector 21422A 2-4 MHz Advance Vision
CA 5-2 21425A 2-5 MHz Advance Vision
Description Part Number Frequency Connector
S4 ultraband sector1 21330A Cannon
S8 ultraband sector 21350A Cannon
c3540 biopsy capable1 21321A 3.5 MHz Cannon
EC6509 endocavity 21336A Cannon
L7535 linear 21359A 7.5 MHz Cannon
L5035 linear 21360A 5 MHz Cannon
TEE (Omniplane II) 21369-60000 6 MHz Cannon
L1038A small parts 21376A Cannon
15-6L intraoperative 21390A Cannon
Pencil 21221-60062 1.9 MHz Pencil
Table 14-1 lists the transducers you can use with the M2540A.
S12 21380A Cannon · Table 14-1 Transducers

Glossary:
2D mode A mode of operation where the display represents reflected ultrasound information as points of
Angio mode This mode presents two-dimensional, real-time Doppler or time shift information superimposed
Color mode This mode presents two-dimensional, real-time Doppler or time shift information superimposed
information, regardless of delay. The detected frequency shifts can produce audio signals, or they
colored pixel information represents the velocities of all the blood flows, or the flow velocities,
as colored pixels on a real-time, gray scale anatomic cross-sectional image. The colored pixel
CW Doppler mode In the Continuous Wave Doppler mode, the transducer array uses separate transmitting and
information represents the amplitudes of the selected blood flows within the angio window.
as colored pixels on a real-time, gray scale anatomic cross-sectional image. Flow directions
toward and away from the transducer are presented as different colors on the display. The
Control Any mechanism (softkey, touch key, push button, toggle, rotary knob, etc.) that changes a
receiving piezoelectric elements. It processes the entire continuum of reflected signal
light of varying intensity arrayed linearly, concordant with the arrival time.
Default preset A preset created at the factory and shipped with the ultrasound system.
parameter that, in turn, affects the displayed image.
can display graphically in a spectral trace.
Control setting The current setting of an active control.
Archive A depository for the final reports.
Custom preset A preset created by the user.
within the color window.

Glossary:
Image An ultrasound view that is either actively displayed or archived. An image can be represented by
Display All the information visible on the system's monitor, including ultrasound images, measurements,
a single frame, a continuous loop, a 3D volume set, a Doppler spectral trace, Doppler audio, an
images, network archive of images, all measurements and calculations, image annotations, and
ECG Electrocardiogram. The system displays a live ECG waveform if one is connected and activated.
reverse video (light-on-dark, or dark-on-light, whichever is the reverse of the normal display).
Live mode Any unfrozen system mode that shows acoustic data in real-time. Live mode includes scanning
Keys on the keyboard are highlighted by lighting an adjacent LED. Softkeys are highlighted by
Live imaging mode Any unfrozen system mode that includes a two-dimensional image being created in real-time
Documentation Everything produced as a result of an ultrasound exam: prints, film, VCR tape, M/O disk of
Highlight An indication that a control is active. Text on a CRT screen is highlighted by displaying it in
Imaging mode 2D, Color, Angio, Mmode, PW Doppler, CW Doppler, 3D, etc.
(that is, not retrieved from VCR tape, Cine Loop memory, etc.).
with a non-imaging transducer. Compare to Live Imaging mode.
the sonographer's or physician's exam-related comments.
tion system The computer databases and networks of an institution.
changing the value associated with them.
system state, patient information, etc.
Institution A hospital, lab, clinic, or office.
ECG waveform, etc.
Hospital informa-

Glossary:
Mmode Method of image processing in which the display is recorded as a spot of light for each detected
A mode that uses pulsed ultrasound. The detected frequency shifts can produce audio signals, or
Sonographer A non-physician who has been trained to perform ultrasound studies. The sonographer performs
sound pulse returning from a moving target. Produces a single-dimensional time/motion display
Scanning window An area of the 2D image pertaining to one of three specific modes: 2D, Color, or Zoom. The
the patient study directly on the patient and captures results for later review by a reviewing
Setup parameters System settings that can be configured to accommodate the user's requirements and
Peripherals Output devices such as VCR, MOD, hard-copy printers, network interface, etc.
windows pertaining to these areas are called the 2D, Flow, and Zoom windows
Preset A group of control settings that optimizes the system for a particular exam.
System text Text showing settings of parameters having to do with system operation.
System settings Refers to the values of a set of controls of the ultrasound system.
Select To indicate a choice, as in choosing an item from a menu.
MOD Magneto-optical disk; a type of removable data storage.
Option A piece of system functionality that a user can purchase.
Modality work list A list of patient names and associated studies to occur.
they can display graphically as a spectral trace.
of detected anatomical structures.
preferences.
physician.
Doppler mode · Pulsed Wave

Glossary:
Transducer icon A graphic symbol that represents the ultrasonic transducer. This symbol indicates the scan plane
and flow imaging using pulse-echo sound waves that are greater than 20 kHz (above the range of
Transducer The element of the ultrasound system that the sonographer places on the patient. Sometimes
called a probe. The transducer emits high-frequency sound waves into the patient and receives
their echoes. The ultrasound system controls the transmitting and receiving of the ultrasound
Ultrasound system Any system that can perform diagnostic sonography, which is medical cross-sectional anatomic
processes the received echoes to construct the displayed images, also in accordance with the
waves according to imaging parameters set by the sonographer. The ultrasound system
by placing it in the proper position and orientation to a body mark icon.
imaging parameters set by the sonographer.
human hearing).