Table of Contents Parts of the CM12 4 The Microscope The Control Panels The Specimen Holder The Side Entry Specimen Chamber Start-up Procedure 7 Powering up the Microscope Saturating the Filament Column Alignment 9 Centering the Beam Alignment of Thermionic Gun Condenser Aperture Adjustment Condenser Astigmatism Correction Sample Loading and Unloading 11 Mounting the Specimen Inserting the Specimen Holder Sample Image Adjustment Optimizing Eucentric Height Adjusting Rotation Center Centering the Pivot Point Removing the Sample Obtaining the Image 14 Fine Tuning the Microscope Capturing the Image Power Down Sequence 15 Photo Credits 16

Figure 1 The Philips CM12 (S)TEM Microscope 1 Electron Gun A filament tip (LaB6 in this case) that emits electrons upon application of electrical current. 2 Condenser Aperture Assembly
An area that houses the condenser aperture, control knobs, and related circuitry. 3 Specimen Chamber/Holder High vacuum tunnel where the specimen is inserted. 4 Objective Aperture Assembly
An area that houses the objective aperture, control knobs, and related circuitry. 5 Field Limiting Aperture Assembly
An area that houses the intermediate lens aperture, control knobs, and related circuitry. 6 Binoculars Eyepieces used to view the projected image up close. 7 Viewing Window Glass windows (another on the left side) where the large phosphor screen can be viewed. 8 Mechanical Stage Controls Two knobs that control the translational coordinates of the sample image.

The Control Panels Left Control Panel
Figure 2 The Control Panel located to the left of the microscope column.
Figure 3 The main control panel on the right hand side of the microscope.
9 Intensity Knob 16 Emission Meter 10 Fine Adjustment Knob 17 High Tension Button 11 Screen 18 Filament Control Knob 12 Screen Control Buttons 19 Ready Button 13 Power Switch 20 Focus/Focus Step Knob 14 Magnification Knob 21 Multifunction Control Knob 15 Beam Shift Knobs
Right Control Panel

Figure 4 The CM12 Specimen Holder 22 Black Cap 24 Specimen Cavity & Clamp 23 O-ring The Side Entry Specimen Chamber

Figure 5 The Side Entry Chamber 25 Tilt Knob 27 Z-Adjustment Knob 26 Indicator Light 28 Tilt Lock Lever Startup Procedure

Powering Up the Microscope 1. Start by logging in to the computer connected to the TEM. Login is a necessary step to prevent unauthorized users from accessing the microscope. Without login, the control panel is not accessible. 2. Turn on the control panel by pushing the Power Switch [13] on the left hand side of the screen.

Figure 6 The main menu screen of the CM12 TEM

Figure 7 The needle will kick twice to the right after the High Tension button is turned on.
Saturating the Filament 1. Once the microscope is powered up, the filament must be heated to adjust the intensity of the electron beam. 2. Go to MODE > CONFIGURATION to view the filament saturation window. The screen will show a FIL LIMIT of 24 indicating maximum saturation.


Column Alignment Centering the Beam 1. To center the beam more easily, adjust the intensity knob [9] to find the cross over i.e. the brightest and more focused beam setting. (Note: if the intensity knob is a little hard to control, press the FINE button [10] to spread or concentrate the beam more slowly.) 2. Check to see if the center of the beam coincides with the black dot on the screen. 3. Use the X & Y shift knobs [15] to center the beam. Alignment of the Thermionic Gun 1. Adjusting the tilt of the gun ensures that the beam of electrons is directed to the center of the sample area. Start by centering the cross-over beam as described earlier.

Figure 9 The shadow of the filament becomes visible at low saturation.
Figure 10 The STIG, DF, and ALGN buttons activate the multi-function knobs.

Condenser Aperture Adjustment Figure 11 A non-concentric beam is seen when the condenser aperture is not aligned.


1. Samples for the CM12 are deposited as thin films on a copper mesh grid such as in Figure 13. This grid is the one loaded into the specimen holder (Figure 4) by first lifting the specimen clamp [24] using a pin-looking tool. (Note: Care must be taken to lift the clamp very gently so as not to break the fragile contraption. In addition, never touch the rod between the o-ring [23] and the specimen cavity at any point during sample mounting.)
Figure 13 An example of a TEM grid used to hold thin film samples
2. Using a pair of tweezers, carefully grab one side of the sample grid and lower it into the specimen cavity. (Note: Never grab the mesh in any other way so as not to distort the mesh or scrape away the sample.) 3. Gently lower the clamp in place using the pin tool to ensure the sample is locked in place. Inserting the Specimen Holder 1. Before inserting the specimen holder into the side entry chamber [3] (also called goniometer), check to make sure the tilt [25] is at 0 degrees and the lever [28] is in lock position. 2. Grab the end of the specimen holder nearest the black cap [22] and ensure that the slit on the cap is at the 8:00 position when inserted to the chamber. Push the holder further until a click as heard. 3. The red indicator light [26] will turn on as the chamber begins to be evacuated. Wait for this light to turn off before proceeding to the next step. 4. When the light goes off, turn the specimen holder counter clockwise until the slit is at the 3:00 position. 5. Gently support the rod as it gets sucked further into the chamber to prevent damage to the tip with the sample. 6. To “see” the sample on the large screen, spread the beam then adjust the stage controls [8] so that the viewing area is at the center of the opaque grid.


Figure 14 The center of the four holes in this sample can be chosen as a reference point for z-height adjustment.
2. Unlock the tilt lever and tilt the sample from +10 degrees to -10 degrees to see how far the reference point moves. Return the sample to 0 tilt. 3. Turn the z-height knob [27] to one direction then tilt the sample +/-10 degrees again to check the movement of the reference point. If the movement becomes smaller, continue turning the knob in that direction. Otherwise, turn the other way. 4. Continue alternating between turning the z-knob and tilting the device until the movement of the reference point reaches a minimum. At this point, the optimal z- height has been reached. 5. Return the tilt to 0 degree then put the lever in lock position. Sample Image Adjustment-Adjusting Rotation Center 1. As in z-height adjustment, begin by choosing a central feature or the sample as a reference point. 2. Press the ALGN button on the control panel and select ROT CENTER from the onscreen menu. 3. The image should wobble in and out i.e. symmetrically around the center when the objective rotation center is aligned. If the reference point moves in addition to the wobbling, adjust the multifunction X/Y knobs until the reference point stays put. 4. Another way to check if the rotation center is aligned is to got to different focus steps by turning the outer knob of the focus-focus step knob [20]. The wobbling of the sample should appear relatively the same at different focus steps.


Figure 15 If the beam splits into two when entering the pivot point alignment menu, the pivot point must be adjusted.
3. Next, select PIVOT PT Y and again coincide the beams if necessary using the multifunction knobs. 4. Press the ALGN button again to exit the align mode. Removing the Sample 1. Ensure once again that the tilt of the sample is at 0 and the level is in lock position. 2. Pull the black cap of the sample holder until a natural stop is felt. 3. Rotate the specimen holder clockwise until the slit returns to the 8:00 position. 4. Gently pull the specimen holder until it fully comes out of the specimen chamber. 5. Remove the sample grid from the holder following the steps for mounting in reverse order.

Obtaining the Image Fine Tuning the Microscope 1. Use the mechanical stage controls to select certain areas of the sample to be captured. 2. Increase the magnification and adjust the focus knobs as necessary to see textures and fine features of the sample. 3. Use the binoculars [6] and small viewing screen as necessary to look at desired areas of the sample. Capturing the Image 1. Spread the beam to illuminate a wide area and take out the viewing screen out of the way of the camera by pulling a lever on the left side of the viewing chamber. 2. Switch to the computer and access the iTEM imaging software. Click on the ACQUIRE button to see a live image of the sample on the computer screen. 3. If ready, click on SNAPSHOT to capture a still image of the sample. 4. Click on STOP ACQUISITION right away to prevent the camera from being exposed to the electron beams for a long period of time. 5. Click on CAMERA CONTROL to lower exposure times when necessary.
Figure 16 The image captured from a sample during an EMS230 lab session

Power Down Sequence 1. Return the viewing screen back to its original position and spread the beam. 2. If high magnification was used, lower the magnification to about 7100x by moving the magnification knob counter-clockwise. 3. Go to MODE > CONFIGURATION then turn the filament knob counter-clockwise until the filament level goes back to zero. (Note: There is no wait time when decreasing the filament saturation. 4. Turn off the HIGH TENSION switch and turn off the control panel. If another user will use the microscope soon after, just press VACUUM from the main menu to see the vacuum status then leave it at that. 5. Cover the viewing windows with their lids. 6. Logoff from the computer to lock the control panel. 7. Sign the microscope user logsheet and indicate if the microscope has been in good working condition.

Figure # Source 1 Mehraeen S. (2007, January 19). 230L, first report. Email attachment. 2 Browning, Nigel (2006). 1Experiment. p. 4, EMS 230L Internal Presentation. 3 Ibid. 4 Ibid. 5 Mehraeen S. (2007, January 19). 230L, first report. Email attachment. 6 Author’s photo 7 Ibid. 8 Ibid. 9 Williams, David B. and Carter, C. Barry (1996). Transmission Electron Microscopy. (Volume I, p. 74). New York: Plenum Press. 10 Author’s photo 11 Williams, p. 137. 12 Ibid. 13 Gilder Grids. Retrieved 2007, January 17 from http://www.gildergrids.co.uk/ 14 Untitled. Retrieved 2007, January 22 from http://www.kwc.edu/academic/biology/TEMpics/teminstallation.htm 15 Author’s photo 16 Ibid.