Instruction/ maintenance manual of the product HH508 Omega Vehicle Security
Go to page of 20
OMEGA HH508/HH509/HH509R Digital Thermometer.
OMEGAnet SM On-Line Service http://www.omega.com Internet e-mail info@omega.com Servicing Nor th America: USA: ISO 9001 Certified Canada: One Omega Drive, Box 4047 976 Bergar Stamford, CT 06907-0047 Laval (Quebec) H7L5A1 Tel: (203) 359-1660 Tel: (514) 856-6928 FAX: (203)359-7700 FAX: (514) 856-6886 e-mail: info@omega.
1 The symbol on the instrument indicates that the operator must refer to an explanation in this manual. WARNING To avoid electrical shock, do not use this instrument when working voltages at the measurement surface over 24V AC or DC. WARNING To avoid damage or burns, do not make temperature measurement in microwave ovens.
2 SPECIFICATIONS ENVIRONMENTAL Ambient Operating Ranges: 0°C to 50°C (32°F to 122°F) <80% R.H. Storage Temperature: -20°C to 60°C (-4°F to 140°F) <70% R.H. GENERAL Display: 3½ digit liquid crystal display (LCD) with a maximum reading of 1999.
3 ELECTRICAL Temperature Scale: Celsius or Fahrenheit user-selectable. Measurement Range: Thermocouple Range K-TYPE(0.1°C) -200°C to 1372°C, -328°F to 1999°F J-TYPE(0.1°C) -210°C to 1200°C, -346°F to 1999°F T-TYPE(0.1°C) -200°C to 400°C, -328°F to 752°F E-TYPE(0.
4 OPERATING INSTRUCTIONS Power Switch The key turns the thermometer on or off. In the MIN MAX record mode can not power off, must leave MIN MAX record mode then power off. °C °F Selecting the Temperature Scale Readings is displayed in either degrees Celsius(°C) or degrees Fahrenheit(°F).
5 MIN MAX record Mode Press MIN MAX key to enter the MIN MAX Recording mode, (displays the Maximum reading, Minimum reading, "MAX-MIN" reading and Average reading stored in record mode). In the this mode the automatic power-off feature is disabled and key and all function key are disabled.
6 TIMER STOP-WATCH Mode (for HH501AR / HH501AS) Press TIMER STOP-W key to enter stop-watch mode stop-watch is displayed. Press START/STOP key to toggle stop-watch starts and stops (Time goes up to 19 minutes & 59.9 seconds). Press CLEAR key to display 0.
7 SET mode (Relative value set and Hi/Lo Limits value set) 1. Press SET key to enter Relative valuse SET mode (Press ENTER key can escape relative valuse set mode), REL set mode. _ = = =.= is displayed (the ".=" is displayed when you set up the fourth umber).
8 Hi/Lo LIMITS mode Press Hi/Lo LIMITS key to enter the Hi/Lo Limits comparative mode, "LIMIT" is displayed. In the this mode the automatic power-off feature is disabled and other funtion key are disabled, when input temperature value exceed Hi.
9 OPERATOR MAINTENANCE WARNING To avoid possible electrical shock, disconnect the thermocouple connectors from the thermometer before removing the cover. Battery Replacement Power is supplied by a 9 volt "transistor" battery. (NEDA 1604, IEC 6F22).
10 RS-232 Operation Using Quick BASIC The following example shows how to send command instruction and receive data responses over the RS-232 interface using Quick BASIC.
11 IF C$ = "-" THEN DEC = DEC * -1 END IF B$ = MID$(A$, 10, 6) DEC1 = 0 N = 6 FOR I = 0 TO 5 C$ = MID$(B$, N, 1) N = N - 1 D = INSTR(F$, C$) - 1 DEC1 = DEC1 + (16 ^ I) * D NEXT DEC1 = DEC1 /.
12 RS-232 Transmit and Receive Comammand Summary 1. Control letter should use Capital Letter. Baud rate: 1200 Parity check: EVEN Data bits: 7 Stop bits: 1 "A" Activate data transmission &quo.
13 08 — Thermocouple Type "K" ® K type "J" ® J type "T" ® T type "E" ® E type "S" ® S type OUTPUT DATA FORMATS The data format consists of 32 bytes.
14 16 — none — " - " 17 — MSD 18 — LSD 19 — MSD 20 — LSD 21 — MSD 22 — LSD (H)ours (M)inutes 23 — Record mode "R" ® Record mode "M" ® MAX mode " .
15 COMMUNICATION 508BR RS-232C "COM1" PORT 4612-IR232 RS232C CABLE 1. Install the RSR232C cable between computer "COM1" and Thermometer RS232C port.
1 HH501(A,B) K / HH501(A,B) J / HH501(A, B)T / HH501( A,B)E HH508BR/HH508/HH509/ HH509R CA LIBRA TION PROCEDURE Note: The following calibration procedure s hould perform only by qualified technicians who have access to the items as fol lowing. Equipment: The class of calibrator ha d better 10 times g reater than the measur ed meter.
W ARRANTY / DISCLAIMER OMEGA ENGINEERING, INC. warr ants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase. OMEGA Warr anty adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time.
M-3203/0799 Where Do I Find Everything I Need for Process Measurement and Control? OMEGA...Of Course! TEMPERATURE þ Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies þ W.
An important point after buying a device Omega Vehicle Security HH508 (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Omega Vehicle Security HH508 yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Omega Vehicle Security HH508 - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Omega Vehicle Security HH508 you will learn all the available features of the product, as well as information on its operation. The information that you get Omega Vehicle Security HH508 will certainly help you make a decision on the purchase.
If you already are a holder of Omega Vehicle Security HH508, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Omega Vehicle Security HH508.
However, one of the most important roles played by the user manual is to help in solving problems with Omega Vehicle Security HH508. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Omega Vehicle Security HH508 along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center