Omega RD200

134 pages·2.6 MB·Sensor·PDF·ENDEFRITNLJA
Omega RD200
Table of Contents 65 sections Preface: Request and notice · Other instruction manuals to be provided separately · FOR SAFE USE · 1. Preconditions for Use · … ShowHide
Preface: Request and noticep.5Other instruction manuals to be provided separatelyp.5FOR SAFE USEp.81. Preconditions for Usep.8WARNINGSp.9L Np.9MAJOR FEATURES AND FUNCTIONSp.10RD200p.11RD2800p.111. GENERAL 1.2 Accessories and Consumablesp.121 Accessories attachedp.122. INSTALLATION 2.1 Location and External Dimensions Locationp.13External Dimensionsp.132. INSTALLATION 2.2 Installation to a Panelp.14Warningp.14RD200 Panel cutout sizep.14RD2800 Panel cutout sizep.14Installation (1) Insert your RD200 (pen type) into the panelp.14Mounting methodp.143. CONFIGULATION 3.1 Frontp.153. CONFIGULATION 3.2 Displayp.16CH 1 CH 2 CH 4 CH 3p.16CH 1 CH 2 CH 3 CH 4p.164. CONNECTIONS 4.1 Terminal Board Arrangementp.174. CONNECTIONS 4.3 Power Terminalsp.19L/N indication of power terminalsp.19LNp.19Be careful with the power voltage and noises. Cautionp.19mark at power terminalsp.19Allowable input voltagep.204. CONNECTIONS 4.5 Alarm Output Terminalsp.21100 VAC 0.5 A 0.2 Ap.21240 V AC 0.2 A 0.1 Ap.21100 V DC 0.3 A 0.1 Ap.214. CONNECTIONS 4.6 Remote Contacts Terminalsp.22No-voltage contacts Warningp.22Remote contacts terminalsp.22COM EX1 EX2 EX3 EX4p.224. CONNECTIONS 4.7 Communications terminalsp.235. INSTALLATION 5.1 Chart Paper Loading (RD200)p.25FEEDp.265. INSTALLATION 5.2 Chart Paper Loading (RD2800)p.27○ ×p.285. INSTALLATION 5.3 Recording Pen Loadingp.29ENTRY REC ON/OFFp.29REC ON/OFFp.296. BASIC OPERATION 6.1 Power Supply and Operationp.31(250V 3A)p.31CH 1 CH 2 CH 3p.316. BASIC OPERATION 6.2 Printing ON/OFF and chart end detectionp.32REC ON/OFF ENTRYp.32ENTRY REC ON/OFF →p.32REC ON/OFF ENTRY →p.323 Chart End Detectionp.326. BASIC OPERATION 6.3 Fast Feeding of Chart Paperp.336. BASIC OPERATION 6.4 Switching Operation Screen (RD200)p.34DISP DISP DISPp.346. BASIC OPERATION 6.5 Switching Operation Screen (RD2800)p.35DISP DISPp.357. PROGRAMMING 7.1 Keys and Charactersp.361 Keys and Their Abbreviationsp.36REC ON/OFF FEED DATA PRINT DISPp.36SPACE COPY(=)p.367 RANGEp.368 ALARMp.36
Page 1 of the Omega RD200 user manual
Page 1 · click to enlarge
Page 1
Omega RD200 diagram (page 1)
Fig. 1
Omega RD200 diagram (page 1)
Fig. 2
Omega RD200 diagram (page 1)
Fig. 3
Omega RD200 diagram (page 1)
Fig. 4
Omega RD200 diagram (page 1)
Fig. 5
Omega RD200 diagram (page 1)
Fig. 6
Omega RD200 diagram (page 1)
Fig. 7
Omega RD200 diagram (page 1)
Fig. 8
Omega RD200 diagram (page 1)
Fig. 9
Omega RD200 diagram (page 1)
Fig. 10
Omega RD200 diagram (page 1)
Fig. 11
Omega RD200 diagram (page 1)
Fig. 12
Omega RD200 diagram (page 1)
Fig. 13
Omega RD200 diagram (page 1)
Fig. 14
Omega RD200 diagram (page 1)
Fig. 15
Omega RD200 diagram (page 1)
Fig. 16
Omega RD200 diagram (page 1)
Fig. 17
Omega RD200 diagram (page 1)
Fig. 18
Omega RD200 diagram (page 1)
Fig. 19
Omega RD200 diagram (page 1)
Fig. 20
Omega RD200 diagram (page 1)
Fig. 21
Omega RD200 diagram (page 1)
Fig. 22
Omega RD200 diagram (page 1)
Fig. 23
Omega RD200 diagram (page 1)
Fig. 24
Omega RD200 diagram (page 1)
Fig. 25
Omega RD200 diagram (page 1)
Fig. 26
Omega RD200 diagram (page 1)
Fig. 27
Omega RD200 diagram (page 1)
Fig. 28
Omega RD200 diagram (page 1)
Fig. 29
Omega RD200 diagram (page 1)
Fig. 30
Omega RD200 diagram (page 1)
Fig. 31
Omega RD200 diagram (page 1)
Fig. 32
Page 2 of the Omega RD200 user manual
Page 2 · click to enlarge
Page 2
Omega RD200 diagram (page 2)
Fig. 33
Omega RD200 diagram (page 2)
Fig. 34
Omega RD200 diagram (page 2)
Fig. 35
Omega RD200 diagram (page 2)
Fig. 36
Omega RD200 diagram (page 2)
Fig. 37
Omega RD200 diagram (page 2)
Fig. 38
Omega RD200 diagram (page 2)
Fig. 39
Page 3 of the Omega RD200 user manual
Page 3 · click to enlarge
Page 3
Omega RD200 diagram (page 3)
Fig. 40
Omega RD200 diagram (page 3)
Fig. 41
Omega RD200 diagram (page 3)
Fig. 42
Omega RD200 diagram (page 3)
Fig. 43
Omega RD200 diagram (page 3)
Fig. 44
Omega RD200 diagram (page 3)
Fig. 45
Omega RD200 diagram (page 3)
Fig. 46
Page 4 of the Omega RD200 user manual
Page 4 · click to enlarge
Page 4
Omega RD200 diagram (page 4)
Fig. 47
Omega RD200 diagram (page 4)
Fig. 48
Omega RD200 diagram (page 4)
Fig. 49
Omega RD200 diagram (page 4)
Fig. 50
Omega RD200 diagram (page 4)
Fig. 51
Omega RD200 diagram (page 4)
Fig. 52
Page 5 of the Omega RD200 user manual — Other instruction manuals to be provided separately
Page 5 · click to enlarge
Page 5

Preface: Request and notice

This instruction describes pen type RD200 series hybrid recorder (100 mm printing width) and pen type

RD2800 series hybrid recorder (180 mm printing width).

In order to use your recorder correctly and safely and to prevent malfunctions, please read this instruction

manual carefully.

Other instruction manuals to be provided separately

This instruction manual describes the operation under the standard specifications and also operations for the optional specifications of (1) alarm output/remote contacts and (2) printing format. For the communications interface unit and the optional specification of mathematical function/totalization, exclusive manuals are provided separately. Also for other optional specifications, of which description of operation is necessary, the relevant instruction manuals are provided respectively. Please read these together with this instruction manual if necessary. Requests

1. To agents or distributors Make sure to pass this instruction manual to final customers. 2. To our valuable customers Keep this instruction manual until disposing of your recorder.

Notices 1. All or any part of this manual may not be duplicated or reproduced in any form, without first obtaining the permission of CHINO. 2. The details of this manual may be subject to change without notice. 3. The contents in this instruction manual have been carefully checked. However, if any question should still arise or if any error, omission or other deficiency be found, please inform your local CHINO sales agent of the details. 4. CHINO will not be responsible for any troubles resulting from the operations of your recorder.

-1-

Page 7 of the Omega RD200 user manual — Table of Contents
Page 7 · Contents
Contents

Contents

Preface: Request and noticep.5Other instruction manuals to be provided separatelyp.5FOR SAFE USEp.81. Preconditions for Usep.8WARNINGSp.9L Np.9MAJOR FEATURES AND FUNCTIONSp.10RD200p.11RD2800p.111. GENERAL 1.2 Accessories and Consumablesp.121 Accessories attachedp.122. INSTALLATION 2.1 Location and External Dimensions Locationp.13External Dimensionsp.132. INSTALLATION 2.2 Installation to a Panelp.14Warningp.14RD200 Panel cutout sizep.14RD2800 Panel cutout sizep.14Installation (1) Insert your RD200 (pen type) into the panelp.14Mounting methodp.143. CONFIGULATION 3.1 Frontp.153. CONFIGULATION 3.2 Displayp.16CH 1 CH 2 CH 4 CH 3p.16CH 1 CH 2 CH 3 CH 4p.164. CONNECTIONS 4.1 Terminal Board Arrangementp.174. CONNECTIONS 4.3 Power Terminalsp.19L/N indication of power terminalsp.19LNp.19Be careful with the power voltage and noises. Cautionp.19mark at power terminalsp.19Allowable input voltagep.204. CONNECTIONS 4.5 Alarm Output Terminalsp.21100 VAC 0.5 A 0.2 Ap.21240 V AC 0.2 A 0.1 Ap.21100 V DC 0.3 A 0.1 Ap.214. CONNECTIONS 4.6 Remote Contacts Terminalsp.22No-voltage contacts Warningp.22Remote contacts terminalsp.22COM EX1 EX2 EX3 EX4p.224. CONNECTIONS 4.7 Communications terminalsp.235. INSTALLATION 5.1 Chart Paper Loading (RD200)p.25FEEDp.265. INSTALLATION 5.2 Chart Paper Loading (RD2800)p.27○ ×p.285. INSTALLATION 5.3 Recording Pen Loadingp.29ENTRY REC ON/OFFp.29REC ON/OFFp.296. BASIC OPERATION 6.1 Power Supply and Operationp.31(250V 3A)p.31CH 1 CH 2 CH 3p.316. BASIC OPERATION 6.2 Printing ON/OFF and chart end detectionp.32REC ON/OFF ENTRYp.32ENTRY REC ON/OFF →p.32REC ON/OFF ENTRY →p.323 Chart End Detectionp.326. BASIC OPERATION 6.3 Fast Feeding of Chart Paperp.336. BASIC OPERATION 6.4 Switching Operation Screen (RD200)p.34DISP DISP DISPp.346. BASIC OPERATION 6.5 Switching Operation Screen (RD2800)p.35DISP DISPp.357. PROGRAMMING 7.1 Keys and Charactersp.361 Keys and Their Abbreviationsp.36REC ON/OFF FEED DATA PRINT DISPp.36SPACE COPY(=)p.367 RANGEp.368 ALARMp.36
Page 8 of the Omega RD200 user manual — 1. Preconditions for Use
Page 8 · click to enlarge
Page 8

FOR SAFE USE

1. Preconditions for Use

Your recorder is designed for indoor use by mounting it on an indoor instrumentation panel. (exception: portable types)

International safety standards … The alarm output (option) with the “c” contact mechanical relay does

  • not conform to the following standards.
  • IEC standards Conforms to safety class I (with PROTECTIVE CONDUCTOR TERMINAL) and IEC1010-1 (OVERVOLTAGE CATEGORY II, POLLUTION DEGREE 2)
  • Enclosure productivity Conforms to IEC529 (IP54)
  • CE (EC Directive) EMC : Conforms to EN61326 Safety : Conforms to EN61010-1 +A2
  • UL standards UL3111-1 (Approval pending)
  • CSA (C-UL) standards CSA C22.2 No. 1010 (Approval pending) 2. Labels on This Instrument

The following labels are used for safe use.

LabelNameMeaning Alert symbol markCaution on handling for prevention of an electric shock, injury or other accidents. Protective conductor terminalA terminal is provided for connection to the protective conductor of the power supply facility in order to prevent any electric shocks.
Omega RD200 — 1. Preconditions for Use diagram (page 8)
Fig. 53 · 1. Preconditions for Use
SymbolsMeaning of symbols This symbol is shown together with relevant titles (or items) where W a r n i n g or Caution is attached. Read them with care. WarningThis symbol indicates a description of cautions for avoiding the possibility of causing serious injury or death due to an electric shock or other accident. CautionThis symbol indicates a description of cautions for avoiding the possibility of causing slight injury to a person or damage to your recorder or to peripheral units. RemarksThis symbol shows a caution when your recorder does not function as specified or when such a possibility exists. ReferenceThis reference serves to indicate supplementary information for handling and operation for your convenience.

-4-

Page 9 of the Omega RD200 user manual — L N
Page 9 · click to enlarge
Page 9

WARNINGS

This paragraph covers important warnings for safety to be observed before reading the instructions. A full understanding of the following warnings is required. These warnings are important for the prevention of danger to human bodies as well as accidents with your recorder.

1. Switch and overcurrent protective device Your recorder is not provided with a power switch or a replaceable overcurrent protective device. Prepare a switch and an overcurrent protective device (circuit breakers, circuit protectors or similar units) for the power supply within 3 m of your recorder in a location where you can reach easily. Use a switch and an overcurrent protective device conforming to IEC948-1 and IEC947-3. 2. Be sure to ground your recorder Before turning on the power, connect the protective conductor terminal your recorder to the protective conductor of the power supply facility. Do not disconnect this wiring in running of your recorder to prevent an electric shock. 3. Before turning on the power supply In order to ensure safety, before turning on the external power switch, make sure that the power voltage is within the range indicated on the power supply label. 4. Don’t repair or modify your recorder. Make sure that a service engineer approved by the CHINO CORPORATION is ONLY permitted to repair or modify your recorder by replacing parts. Otherwise, it may be damaged or will not function correctly, or an accident such as an electric shock may result. For usual operation, it is not necessary to pull out an internal unit from a case.

The following fuse is mounted in the power supply unit of your recorder for safety use. However, this fuse is not replaceable.

L N

Omega RD200 — L N diagram (page 9)
Fig. 54

Overcurrent protective device (250V 3A)

To the protective conductor of power supply facility

Fuse in power supply unit Reference

ManufacturerModelRatings SCHURTERSPT001.2508250VAC 2.5A LITTEL FUSE21502.5 WICKMANN19181
Omega RD200 — L N diagram (page 9)
Fig. 55

100-240VAC 50/60Hz 60VA MAX

5. Use your recorder in accordance with this instruction manual.

Use your recorder correctly and safely in accordance with this instruction manual. CHINO CORPORATION will

not be responsible for any injury, damage, lost profit or any other claim, which may result from its wrong use. 6.Turn off the power supply if an abnormal symptom occurs. If any abnormal odor, noise or any smoke occurs, or if your recorder becomes too hot to be touched, turn off the power supply immediately and contact your local CHINO sales agent.

Show 5 more labels / short text on this page
Power/protective conductor terminals · Power supply · Switch · Power supply label · -5-
Page 10 of the Omega RD200 user manual — MAJOR FEATURES AND FUNCTIONS
Page 10 · click to enlarge
Page 10

MAJOR FEATURES AND FUNCTIONS

Your recorder can record temperature and other various industrial variables on a 100 mm (if your recorder is RD200) or 180 mm (if your recorder is RD2800) chart for 1 to 4 channels (depending on the number of pens).

(1) Trace printing by pens (2) Digital data printing to print measured values and other data 1 Features

  • Major features are shown below.
  • Universal input. A range can be selected optionally for every channel from 10 DC voltage ranges, 36 thermocouple ranges and 11 resistance thermometer ranges.
  • Universal power supply. The working voltage range is 100 to 240 V AC, 50/60 Hz.
  • International safety standards…. CE: Conformance, UL and CSA (C-UL): Approval pending
  • The basic operation should be carried out after programming range numbers and the trace printing range.
  • You can execute all operation by the front keyboard without pulling out the internal unit. The internal unit cannot be pulled out. 2 Functions

Major functions are shown below.

DisplayPrinting ● Simultaneous display of the measured values for 1 to 4 channels. ● Analog indication of measured values for 1 to 4 channels with bargraphs. ● Six status lamps ● Switching the operation screens (Measured value, Clock and Alarm Activation) ●Measured values blinks when alarm activates. (Note)● Trend tracing for 1 to 4 channels ● Fixed-time printing (time line, time, scale, engineering unit, tag) ● Periodic data printing (Measured values are digitally printed at preset intervals.) ● Digital data printing (Measured values are digitally printed on demanding.) ● Message printing ● On or off of time axis synchronization (POC)

(Note) Programming of alarm is necessary. Alarm output is only available when your recorder is with the option

of “Alarm Output”.

-6-

Page 11 of the Omega RD200 user manual — RD200
Page 11 · click to enlarge
Page 11
1. GENERAL 1.1 Confirmation of Model No.

Pen type RD200 series and RD2800 series have various specifications. Confirm Model No. of your recorder. Labels showing Model No. are affixed to the top side of the case and to the inside of the internal unit.

RD200

RD2800

RD204 Model AL00 X A001 Serial No. MADE IN JAPAN

RD2804 Model AH00 X A001 Serial No. MADE IN JAPAN

Omega RD200 — RD200 diagram (page 11)
Fig. 56
Omega RD200 — RD200 diagram (page 11)
Fig. 57
Omega RD200 — RD200 diagram (page 11)
Fig. 58
Omega RD200 — RD200 diagram (page 11)
Fig. 59
Show 5 more labels / short text on this page
Label 1 · Label 1 · Label 2 · Label 2 · -7-
Page 12 of the Omega RD200 user manual — 1 Accessories attached
Page 12 · click to enlarge
Page 12

1. GENERAL 1.2 Accessories and Consumables

1 Accessories attached

NamesQtyRemarks (1) Chart1RD200-CP-0/100 (2) Mounting bracket2For installation to a panel. (3) Channel indicating card1Inserted inside the door. (4) Cartridge pen (Red)1 (5) Cartridge pen (Green)1Only for 2 to 4-pen type (6) Cartridge pen (Blue)1Only for 3 to 4-pen type (7) Cartridge pen (Brown)1Only for 4-pen type (8) Plotter pen1 (9) Terminal screw5Spare (10) Lubricating oil1For the main shaft of the plotter pen (11) Instruction manual1This manual
NamesQtyRemarks (1) Chart1180A-CP-0/100 2) Mounting bracket (left, right)1 setFor installation to a panel. (3) Channel indicating card1Inserted inside the door. (4) Cartridge pen (Red)1 (5) Cartridge pen (Green)1Only for 2 to 4-pen type (6) Cartridge pen (Blue)1Only for 3 to 4-pen type (7) Cartridge pen (Brown)1Only for 4-pen type (8) Plotter pen1 (9) Terminal screw5Spare (10) Lubricating oil1For the main shaft of the plotter pen (11) Instruction manual1This manual (12)Wrench1For tightening mounting bracekts

Separate instruction manuals are attached when your recorder is with the options of "Communications interface" and "Math-function and totalizer”. 2 Consumables

Charts and pens are consumables. For your ordering, refer to the following table.

Omega RD200 — 1 Accessories attached diagram (page 12)
Fig. 60 · 1 Accessories attached
ArticleModelMin. qty Cartridge penRed (1st pen)RD200-013 pieces/ bag for each color Green (2nd pen)RD200-02 Blue (3rd pen)RD200-03 Brown (4th pen)RD200-04 Plotter penPurpleRD200/ RD2800-Plotter3 pieces/ box Chart10 metersRD200-CP-0/10015 charts/ box
ArticleModelMin. qty Cartridge penRed (1st pen)RD2800-013 pieces/ bag for each color Green (2nd pen)RD2800-02 Blue (3rd pen)RD2800-03 Brown (4th pen)RD2800-04 Plotter penPurpleRD200/ RD2800-Plotter3 pieces/ box Chart20 meters180A-CP-0/10015 charts/ box
Show 20 more labels / short text on this page
(1) · (4) – (7) · (8) · (2) · (1) · (2) · (12) · (9) · (3) · (10) · (3) · Front · (9) · Front · Rear · (8) · Rear · (4) – (7) · (10) · -8-
Page 13 of the Omega RD200 user manual — RD2800
Page 13 · click to enlarge
Page 13

2. INSTALLATION 2.1 Location and External Dimensions Location

Install your recorder at the following place so as not to affect the measuring accuracy and recording operation unfavorably. (1) Industrial environment

Install your recorder in a place where no inflammable gas exists and no dust, smoke, vapor, or other dangerous substance exists.

Select a place being separated from electric field and magnetic field generating sources and also free of mechanical vibrations and shocks.

  • Overvoltage Category …………………………. ΙΙ
  • Pollution Degree ………………………………… 2
  • Altitude …………………………… 2000m or lower
  • Working place ……………..………Indoors (3) Ambient temperature and humidity

(4) Mounting angle and display view angle

Make sure not to expose your recorder to direct sunlight and not to closely place other materials to it for preventing rise of its temperature.

  • Lateral tilting……….. 0 ° to 10 °
  • Longitudinal tilting…..Forward tilting: 0 ° Backward tilting: 0 to 30

° • View angle … -10° to +30° based on horizontal

  • The recommended ambient temperature and humidity are about 23 ° C and about 50%RH.
  • Make sure not to expose your recorder to hot air higher than 70 ° C.
  • Make sure not place any heat source near to the terminal board of your recorder.

Approx. 30° View angle

Approx. 10°

External Dimensions

RD200

Omega RD200 — RD200 diagram (page 13)
Fig. 61
Omega RD200 — RD200 diagram (page 13)
Fig. 62

*1: Recorders with the options of MOS relay or ”c” contact alarm output, and communications interface *2: Recorders with the option of “a” contact alarm output

RD2800

Omega RD200 — RD2800 diagram (page 13)
Fig. 63
Omega RD200 — RD2800 diagram (page 13)
Fig. 64

*1: Recorders with the options of MOS relay or ”c” contact alarm output, and communications interface *2: Recorders with the option of “a” contact alarm output

Show 9 more labels / short text on this page
(2) Atmosphere · Display · □137 · 144 24 63 · 260(274)*1 (285)*2 · □277 · 288 297 · 24 220(236)*1 (247)*2 · -9-
Page 14 of the Omega RD200 user manual — RD200 Panel cutout size
Page 14 · click to enlarge
Page 14

2. INSTALLATION 2.2 Installation to a Panel

Install your recorder to a panel.

Warning

Except portable types, your recorder is designed to install to a panel. Use a panel made of a steel plate of 2 mm to 6 mm in thickness.

RD200 Panel cutout size

RD2800 Panel cutout size

Installation (1) Insert your RD200 (pen type) into the panel

Mounting method

  • Minimum clearance for closedinstallation
  • Minimum clearance for closedinstallation

(1) Insert your RD2800 (pen type) into the panel

cutout from the front of the panel. (2) Fix your RD200 (pen type) to the panel by the

cutout from the front of the panel. (2) Gently screw a mounting screw into the screw

mounting brackets (screw tightening torque: 1.0 N•m). Attach 2 mounting brackets to the top and bottom of this instrument.

hole of the case (left, right). (3) Attach the mounting bracket to the case by

putting the mounting screw in the large hole of the bracket. Slide the bracket downward to attach it closely to the panel and then tighten the screw with a wrench (screw tightening torque: 1.2 N•m).

Omega RD200 — RD200 Panel cutout size diagram (page 14)
Fig. 65 · RD200 Panel cutout size
Omega RD200 — RD200 Panel cutout size diagram (page 14)
Fig. 66 · RD200 Panel cutout size
Omega RD200 — RD200 Panel cutout size diagram (page 14)
Fig. 67 · RD200 Panel cutout size

Mounting bracket: Slide it downward and tighten with a wrench.

Mounting bracket Panel thickness

(2 to 6 mm)

Panel thickness (2 to 6 mm)

Omega RD200 — RD200 Panel cutout size diagram (page 14)
Fig. 68 · RD200 Panel cutout size
Omega RD200 — RD200 Panel cutout size diagram (page 14)
Fig. 69 · RD200 Panel cutout size
Show 11 more labels / short text on this page
138 +1 0 · 281 +1 0 · 138 +1 · 281 +1 · (mm) · (mm) · Mounting bracket · Mounting screw · Panel · Mounting bracket · -10-
Page 15 of the Omega RD200 user manual — 3. CONFIGULATION 3.1 Front
Page 15 · click to enlarge
Page 15

3. CONFIGULATION 3.1 Front

All operations of your recorder including the loading of pens can be executed from its front.

Omega RD200 — 3. CONFIGULATION 3.1 Front diagram (page 15)
Fig. 70 · 3. CONFIGULATION 3.1 Front

RD2800 The window of the door is made of glass material. To prevent injuries by destroying it, do not apply an impact or strong force to the door. (1) Display panel

The display panel is consisted of character displays, status lamps and bargraph displays. For loading the pens, open the display panel by swinging it out from the right side. An illumination for chart is mounted on the back of the display panel. (2) Chart paper cassette

The chart paper cassette is for loading a chart to your recorder. Remove it for the loading of chart. When your recorder is RD200, remove the chart paper cassette by swinging down the keyboard. (3) Keyboard

The sheet switch type keys are used in the keyboard. ● Operation keys

Omega RD200 — 3. CONFIGULATION 3.1 Front diagram (page 15)
Fig. 71 · 3. CONFIGULATION 3.1 Front

Switches operation screen. 6.4 + *1 → List printing 9.3

(3 sec. or more) (With recording disabled ) Pen replacing mode 5.3

*1: Selection of List 1 or List 2. *2: Selection of Message No. (1 to 5) ● Programming keys The keys are different on programming items. See Section 7.3 for details.

-11-

Page 16 of the Omega RD200 user manual — CH 1 CH 2 CH 3 CH 4
Page 16 · click to enlarge
Page 16

3. CONFIGULATION 3.2 Display

RD200

CH 1 CH 2 CH 4 CH 3

Omega RD200 — RD200 diagram (page 16)
Fig. 72

CH 1 CH 2 CH 3 CH 4

CH 1 CH 2 CH 3 CH 4 CLOCK CHART SPPED

Omega RD200 — CH 1 CH 2 CH 3 CH 4 diagram (page 16)
Fig. 73

CH 1 CH 2 CH 3 CH 4

Omega RD200 — CH 1 CH 2 CH 3 CH 4 diagram (page 16)
Fig. 74
  • Operation mode: Displays measured value, time, chart speed, and alarm status of each channel (CH). See Section 6.4 for details.
  • Programming mode: Displays programming items and programming parameters in an interactive method. (2) Underline

Shows the trace printing color of each channel (CH). (CH 1: Red, CH 2: Green, CH3: Blue, CH 4: Brown) These underlines also function as a cursor appearing at the digit for programming parameter in the programming mode. (3) Status

Display Lighting condition Section

ALARM When an alarm activates. 10.3 CHART END When the end of chart is detected. 6.2 FAIL When the hardware related to servo-circuit/mechanism is abnormal. POC When the time axis synchronization is enabled. 9.6 RECORD ON When the printing is on (enable). 6.2 KEY LOCK When the keys are locked. 11.12

The bargraphs indicate the measured value of each channel in an analog form. These indications are interlocking to the positions of the pens for trace printing. Resolution RD200: 1/50 (2%), 51 segments

RD28001/100 (1%), 101 segments
Show 9 more labels / short text on this page
(1) (2) · (3) · (4) · RD2800 (1) Characters · (1) (2) · (3) · (4) · (4) Bargraph · -12-
Page 17 of the Omega RD200 user manual — RD200
Page 17 · click to enlarge
Page 17

4. CONNECTIONS 4.1 Terminal Board Arrangement

The terminal boards shown in the following figure are of recorders with the options (alarm output + remote contacts and communications interface).

RD200

RD2800

(option) Power supply terminals

(option) Power supply terminals

Omega RD200 — RD200 diagram (page 17)
Fig. 75
Omega RD200 — RD200 diagram (page 17)
Fig. 76

Remote contacts terminals (option)

Power supply terminals (option)

Remote contacts terminals option)

Power supply terminals (option)

*The mechanical relay “c” contact output is consisted of three terminals of N.O, COM and N.C.

The input terminal and alarm terminal blocks are removable. Reference

The input terminal block and alarm terminal block (including the remote contacts terminal block) are removable for easy connections. Each terminal block can be taken out by removing two mounting screws. Each terminal block is connected to your recorder by a connector. For mounting or dismounting the terminal block, turn off the power switch to prevent the electric circuits from being damaged. RD200 RD2800

Omega RD200 — RD200 diagram (page 17)
Fig. 77
Omega RD200 — RD200 diagram (page 17)
Fig. 78
Omega RD200 — RD200 diagram (page 17)
Fig. 79
Omega RD200 — RD200 diagram (page 17)
Fig. 80

Removed terminals have to be remounted to the recorder to which they were originally mounted. Mounting them to other recorders may cause a malfunction in scaling.

Show 9 more labels / short text on this page
Communications terminals · Communications terminals · Measuring input terminals · Measuring input terminals · Input terminal block · Input terminal block · Mounting screw · Mounting screw · -13-
Page 18 of the Omega RD200 user manual — Warning
Page 18 · click to enlarge
Page 18
4. CONNECTIONS 4.2 Cautions on Connections Observe the following cautions during connections for securing safety and reliability. 1 Power supply

4 Keep the input/output connections away from noise source Keep all connection cables away from noise source as far as possible, otherwise a malfunction may occur. Provide a solution if the cables cannot be separated from a noise source due to unavoidable circumstances.

Use a single-phase power supply having a stable voltage without any waveform distortion for the purpose of preventing wrong operations.

Warning

A switch and an overcurrent protective device

Prepare a switch and an overcurrent protective device (3 A) to the power supply for preventing an accidental electric shock during connection work. This instrument is not provided with any replaceable overcurrent protective device.

Major noise sources Remedial measures

Insert noise filters between power terminals and input/output terminals. A CR filter is often used. 5 Use crimp style terminals

Electromagnetic switch, etc. Power line having waveform distortion, Inverter, Thyristor regulator

Turn off the power supply before starting connections.

Warning

Make sure to turn off the power supply before connecting the power and the input/output terminals to prevent an electric shock. 2 Keep the input/output connections

(1) Mount crimp style terminals for connection

cable terminations to prevent any looseness or disconnection of terminals or a short-circuit failure between terminals. (2) Use the crimp style terminals with an

insulation sleeve to prevent an electric shocks. 6 Unused terminals

away from high voltage power circuits Don’t place the input/output cables close to or in parallel with any strong power circuits including power lines. Place the cables 50cm or more away from high voltage power circuits when they are placed close to or in parallel to other circuits. 3 Keep the thermocouple input away

Don’t use any unused terminals for relaying, otherwise the electric circuits may be damaged.

Omega RD200 — Warning diagram (page 18)
Fig. 81 · Warning

Secure the connected cables so as not to allow them to be hooked by a person or a substance, otherwise the connections may be cut and disrupted, and may cause an electric shock or other accidents.

from a heat source For thermocouple inputs, keep the input terminals away from a heat source (a heating body) to reduce a reference junction compensation error. Don’t expose the input terminals to direct sunlight, etc.

8 or less 3.7 or more 3.7 or more 8 or less

*Use Type O chip (on the left) whenever possible.

-14-

Page 19 of the Omega RD200 user manual — mark at power terminals
Page 19 · click to enlarge
Page 19

4. CONNECTIONS 4.3 Power Terminals

1 Power and protective conductor terminals 2 Connection of power terminals For connection to the power terminals, use a 600 V PVC insulated cable (IEC 227-3 See “Caution”) terminated by the crimp style terminals with insulation sleeve. Note) Use the cords approved by the following standards.

Make sure to turn off the power supply before the connections to the power and protective conductor terminals for preventing an electric shock. This indication conforms to the CSA standard, Canada. The live side of the single-phase AC power supply is indicated as L, and the neutral side is indicated as N. Observe the L and N connections for obtaining satisfactory performance.

Turn off the power supply.

Warning

L N

100-240V AC 50/60Hz 60VA MAX

L/N indication of power terminals

Power supply (voltage, frequency, power consumption)

(1) IEC 227-3 (2) ANSI/UL817, (3) CSA C22.2 No.21/49

LN

Mount the terminal cover after connections. (Illustration below)

Be careful with the power voltage and noises. Caution

Omega RD200 — Be careful with the power voltage and noises. Caution diagram (page 19)
Fig. 82 · Be careful with the power voltage and noises. Cautio…

The power voltage of your recorder is indicated beside the power terminals. Don’t apply any voltage other than the rated voltage. Otherwise a malfunction may result. If noise is contaminated in the power supply, provide a noise reduction transformer, etc. 3 Connection of protective conductor terminal Make sure to connect this terminal to the protective conductor of the power supply facility. For this connection, use a cable terminated by the crimp style terminals with insulation sleeve. • Grounding wire: Copper wire 2 mm2 or more

Omega RD200 — Be careful with the power voltage and noises. Caution diagram (page 19)
Fig. 83 · Be careful with the power voltage and noises. Cautio…
Omega RD200 — Be careful with the power voltage and noises. Caution diagram (page 19)
Fig. 84 · Be careful with the power voltage and noises. Cautio…

A copper wire with diameter of2 mm2 or more (green/yellow)

Omega RD200 — Be careful with the power voltage and noises. Caution diagram (page 19)
Fig. 85 · Be careful with the power voltage and noises. Cautio…

Make sure to connect to the protective conductor of the power supply facility.

Omega RD200 — Be careful with the power voltage and noises. Caution diagram (page 19)
Fig. 86 · Be careful with the power voltage and noises. Cautio…

600 V vinyl insulated cable

Omega RD200 — Be careful with the power voltage and noises. Caution diagram (page 19)
Fig. 87 · Be careful with the power voltage and noises. Cautio…

mark at power terminals

Warning

A voltage of 100 to 240 VAC is applied to the power terminals after connections. Be sure to mount the power terminal cover to prevent an electric shock.

Show 4 more labels / short text on this page
Power terminals · Protective conductor terminal · Power supply · -15-
Page 20 of the Omega RD200 user manual — A
Page 20 · click to enlarge
Page 20
Omega RD200 — Allowable input voltage diagram (page 20)
Fig. 88 · Allowable input voltage

1 Measuring input terminals Make sure to turn off the power supply to prevent an electric shock. (1) Measuring input terminals are located on the

3 Connections of thermocouple inputs Make sure to use thermocouple wires (or extension wires) to the input terminals of your recorder. If a copper wire is used halfway, a noticeable measuring error occurs. Don’t use a pair of thermocouple wires in parallel with other instruments (controller, etc.), otherwise a malfunction may occur. 4 Connections of resistance thermometer inputs Use a 3-core cable where each lead wire has an equal resistance value. Don’t use one resistance thermometer in parallel with other instruments (controller, etc.).

down left of the terminal board. (2) For the connections to the input terminals, use

cables terminated by crimp style terminals with insulation sleeves.

Allowable input voltage

1 2 3 4

Omega RD200 — Allowable input voltage diagram (page 20)
Fig. 89 · Allowable input voltage

Input type Allowable input voltage

±10 VDC (range± 2V or less) ±60 VDC (range: ± 5V or more)
Omega RD200 — Allowable input voltage diagram (page 20)
Fig. 90 · Allowable input voltage

Resistance thermometer input ±6 VDC

Omega RD200 — Allowable input voltage diagram (page 20)
Fig. 91 · Allowable input voltage

2 Connections of DC voltage (current) input Use twisted cables for instrumentation as the input cables for the purpose of suppressing noise. For current inputs, mount shunt resistors (Section 16.2) to the channels to be measured before connections.

Omega RD200 — Allowable input voltage diagram (page 20)
Fig. 92 · Allowable input voltage
Omega RD200 — Allowable input voltage diagram (page 20)
Fig. 93 · Allowable input voltage
Omega RD200 — Allowable input voltage diagram (page 20)
Fig. 94 · Allowable input voltage

1 2 3 4 (+)

Omega RD200 — Allowable input voltage diagram (page 20)
Fig. 95 · Allowable input voltage
Omega RD200 — Allowable input voltage diagram (page 20)
Fig. 96 · Allowable input voltage
Omega RD200 — Allowable input voltage diagram (page 20)
Fig. 97 · Allowable input voltage

Twisted cable for instrumentation

1 2 3 4

3-core cable (Same diameter, same length)

Omega RD200 — Allowable input voltage diagram (page 20)
Fig. 98 · Allowable input voltage

A

B

B

Omega RD200 — A diagram (page 20)
Fig. 99
Omega RD200 — A diagram (page 20)
Fig. 100
Omega RD200 — A diagram (page 20)
Fig. 101

A high voltage may be applied to the measuring input terminals due to common mode noise. The allowable noise value is 30 VAC or less, or 60 VDC or less. Make sure that the noise is lower than the allowable value. Mount the terminal cover after connections for the purpose of preventing an electric shock and to protect the input wires. In the case of thermocouple input, the mounting of the terminal cover can reduce the reference junction compensation error.

Show 10 more labels / short text on this page
Red (+) · Voltage, Thermocouple input · White (-) · Extension wire · Thermocouple · (-) · DC voltage input · Terminal cover · Resistance thermometer · -16-
Page 21 of the Omega RD200 user manual — 240 V AC 0.2 A 0.1 A
Page 21 · click to enlarge
Page 21

4. CONNECTIONS 4.5 Alarm Output Terminals

These terminals are for the alarm output (option), which is (1) MOS relay type, (2) mechanical relay “c” contact type or (3) mechanical relay “a” contact type. The mechanical relay “c” contact type does not conform to the international safety standards.

1 Alarm output terminals The terminal arrangement depends upon the type of alarm output.

Omega RD200 — 4. CONNECTIONS 4.5 Alarm Output Terminals diagram (page 21)
Fig. 102 · 4. CONNECTIONS 4.5 Alarm Output Terminals

Terminal and relay No.

Terminal block RD200 RD2800

1 2 3 4 5 6 1 2 3 4 5 6

1 No. 1 to 6 No. 1 to 6

2 No. 1 to 12

[Remarks]Only one terminal block is provided for the remote contacts terminals.

2 Connections Turn off the power supply and buffer relay power supply before the connections for preventing an electric shock. (1) Connect cables to a load via a buffer relay. (2) Use cables with crimp style terminals and

Omega RD200 — 4. CONNECTIONS 4.5 Alarm Output Terminals diagram (page 21)
Fig. 103 · 4. CONNECTIONS 4.5 Alarm Output Terminals

insulation sleeves for the alarm output terminals.

Omega RD200 — 4. CONNECTIONS 4.5 Alarm Output Terminals diagram (page 21)
Fig. 104 · 4. CONNECTIONS 4.5 Alarm Output Terminals
*Contact protective element (It is recommended to mount this element on the “a” side.)

The alarm output of your recorder may become defective. This may be caused by wrong operation, failure or other abnormal inputs. Take safety measures against an output failure before use if necessary.

* N.C terminalOnly for the mechanical relay “c” contact output

A buffer relay power supply is applied to the alarm output terminals after connections. Don’t touch these terminals for preventing an electric shock. Make sure to mount the terminal cover after connection.

3 Cautions on connections 1) Output contact rating (1) MOS relay

Maximum voltage 240 V (AC, DC)

Maximum current 50 mA (AC, DC)

2. Mounting of contact protective element (1) Mount a contact protective element conforming to the buffer relay. (2) To prevent a malfunction being caused by a

Power supply Resistive load Inductive load

100 VAC 0.5 A 0.2 A

light load, the most effective mounting position for the element is on the coil side of the buffer relay ('a' in the above diagrams) (3) The MOS relay will be broken, even if a

240 V AC 0.2 A 0.1 A

100 V DC 0.3 A 0.1 A

signal exceeding the contact rating is momentarily applied.

(2) Mechanical relay · -17-

Page 22 of the Omega RD200 user manual — COM EX1 EX2 EX3 EX4
Page 22 · click to enlarge
Page 22

4. CONNECTIONS 4.6 Remote Contacts Terminals

These terminals are for the remote contacts (option). For the remote contacts, see Section 13. 1 Remote contacts terminals 2 Wiring To prevent an electric shock, make sure to turn off the power supply before wirings. (1) The signals applied to the remote contacts

No-voltage contacts Warning

Omega RD200 — No-voltage contacts Warning diagram (page 22)
Fig. 105 · No-voltage contacts Warning
  • For the contacts to be connected to the remote contacts terminals, use a switch or relay driven at 30 VAC or lower, or 60 VDC or lower, or manual contacts for very light loads.
  • Voltage when the contact is open: Approx. 5 V
  • Current when the contact is short: Approx. 2 mA 2. Terminal allocation for operation It is required to set operations to be allocated to each terminal (EX1 to EX4). 3. Operations to be set (1) Printing ON/OFF and chart speed

Alarm output terminal (option)

Remote contacts terminals

Omega RD200 — Remote contacts terminals diagram (page 22)
Fig. 106 · Remote contacts terminals

COM EX1 EX2 EX3 EX4

Omega RD200 — COM EX1 EX2 EX3 EX4 diagram (page 22)
Fig. 107

terminals should be non-voltage contact signals. (2) Connect wires to the remote contacts

terminals with crimp style terminals and insulation sleeves 1. Remote contacts operations (1) Printing ON/OFF and chart speed

selection from 3 speeds (Using 2 terminals EX1 and EX2) (2) Message (Nos. 1 to 5) selection and

printing (Using all of 4 terminals EX1 to EX4) (3) Message (Nos. 1 and 2) selection and

selection from 3 speeds

Setting of 3 chart speeds (2) Message selection and printing

printing (Using 2 terminals EX3 and EX4) (4) Digital data printing * (5) List (Nos. 1 to 3) printing * (6) Operation record (Nos. A to D) printing * (7) "Totalizing" (option) reset * * Using any one terminal.

Setting of messages Nos. 1 to 5 (3) Operation recording

Setting of recording positions of operation records Nos. A to D

Remote contacts terminals · Remote contacts Reference · -18-

Page 23 of the Omega RD200 user manual — 4. CONNECTIONS 4.7 Communications terminals
Page 23 · click to enlarge
Page 23

4. CONNECTIONS 4.7 Communications terminals

These terminals are for the communications interface (option). For details of the connection, refer to the instruction manual for the “Communications Interfaces” provided separately. 1 Communications terminals

Omega RD200 — 4. CONNECTIONS 4.7 Communications terminals diagram (page 23)
Fig. 108 · 4. CONNECTIONS 4.7 Communications terminals

RS-232C Terminals RS-422A Terminals RS-485 Terminals 2 RS-232C Connections When your recorder is with the communications interface of RS-232C, three terminals of SD, RD and SG are used but any control signal is not used. General personal computers are controlled by control signal. Wiring processing for control signal in a connector depends upon how the control signal is used in a personal computer. For details, refer to the instruction manual for your personal computer. 1) 9-pin connector

SD RD SG SDA SDB RDA RDB SG SA SB SG

-19-

Page 24 of the Omega RD200 user manual — 4. CONNECTIONS 4.7 Communications terminals
Page 24 · click to enlarge
Page 24
Omega RD200 — 4. CONNECTIONS 4.7 Communications terminals diagram (page 24)
Fig. 109 · 4. CONNECTIONS 4.7 Communications terminals

3 RS-422A, RS-485 Connections RS-422/485 communications interface is connected to a personal computer via a line converter (our Model SC8-10: sold separately). Three signals of SD, RD and SG are used between the line converter and the personal computer but any control signal is not used. Wiring processing for control signal in a connector is necessary in the same as 2 RS-232C connections.

Omega RD200 — 4. CONNECTIONS 4.7 Communications terminals diagram (page 24)
Fig. 110 · 4. CONNECTIONS 4.7 Communications terminals

Within 15 m Within a total length of 1.2 km (max. 31 recorders)

*For the details of the wiring, refer to 2 RS-232C connections.

resistance 100 Ω, 1/4W Example of 9-in connector

Within 15 m Within a total length of 1.2 km (max. 31 recorders)

*For the details of the wiring, refer to 2 RS-232C connections”.

-20-

Page 25 of the Omega RD200 user manual — 5. INSTALLATION 5.1 Chart Paper Loading (RD200)
Page 25 · click to enlarge
Page 25

5. INSTALLATION 5.1 Chart Paper Loading (RD200)

1 Chart Paper Cassette Removal 1) Open the door.

2 Chart Paper Loading 1) Open the chart paper holders. Open the chart paper guide and the stripper plate.

Omega RD200 — 5. INSTALLATION 5.1 Chart Paper Loading (RD200) diagram (page 25)
Fig. 111 · 5. INSTALLATION 5.1 Chart Paper Loading (RD200)
Omega RD200 — 5. INSTALLATION 5.1 Chart Paper Loading (RD200) diagram (page 25)
Fig. 112 · 5. INSTALLATION 5.1 Chart Paper Loading (RD200)
Omega RD200 — 5. INSTALLATION 5.1 Chart Paper Loading (RD200) diagram (page 25)
Fig. 113 · 5. INSTALLATION 5.1 Chart Paper Loading (RD200)
2. Swing down the keyboard.
Omega RD200 — 5. INSTALLATION 5.1 Chart Paper Loading (RD200) diagram (page 25)
Fig. 114 · 5. INSTALLATION 5.1 Chart Paper Loading (RD200)
2. Prepare a chart paper Shuffle both end of the chart paper to prevent two or more chart papers from fed being together.
Omega RD200 — 5. INSTALLATION 5.1 Chart Paper Loading (RD200) diagram (page 25)
Fig. 115 · 5. INSTALLATION 5.1 Chart Paper Loading (RD200)
Omega RD200 — 5. INSTALLATION 5.1 Chart Paper Loading (RD200) diagram (page 25)
Fig. 116 · 5. INSTALLATION 5.1 Chart Paper Loading (RD200)
3. Remove the chart paper cassette . Pens lift up when the chart paper cassette is removed.
3. Put it into the chart paper housing. The shapes of the sprocket holes on the right and left sides are different. The right side holes are an elliptical shape.
Omega RD200 — 5. INSTALLATION 5.1 Chart Paper Loading (RD200) diagram (page 25)
Fig. 117 · 5. INSTALLATION 5.1 Chart Paper Loading (RD200)

Sprocket holes (circular) Printing side

Omega RD200 — 5. INSTALLATION 5.1 Chart Paper Loading (RD200) diagram (page 25)
Fig. 118 · 5. INSTALLATION 5.1 Chart Paper Loading (RD200)
Omega RD200 — 5. INSTALLATION 5.1 Chart Paper Loading (RD200) diagram (page 25)
Fig. 119 · 5. INSTALLATION 5.1 Chart Paper Loading (RD200)
Show 11 more labels / short text on this page
Door · Stripper plate · Chart paper guide · Keyboard · Guide rail · Sprocket holes (elliptical) · Drum · Guide · Grip · Chart paper housing · -21-
Page 26 of the Omega RD200 user manual — FEED
Page 26 · click to enlarge
Page 26

5. INSTALLATION 5.1 Chart Paper Loading (RD200)

3 Chart Paper Setting 1) Draw out the chart paper approx. 20 cm and close the rear stripper plate. 2) Set the holes on the chart paper over the sprockets at both ends of the drum. 3) Use the thumb wheel to advance the chart paper for 2 to 3 folds into the chart tray. 4) Close the front chart paper guide. Make sure that the holes are set over the sprockets.

Turning direction of thumb wheel

Don’t turn the thumb wheel inward. The chart paper cannot be fed backward. This action may cause a chart paper feed failure. 2) Place the chart paper cassette in its original position Place the chart paper cassette with the chart paper loaded in your recorder. (1) Guide rails for the chart paper cassette are

Omega RD200 — 5. INSTALLATION 5.1 Chart Paper Loading (RD200) diagram (page 26)
Fig. 120 · 5. INSTALLATION 5.1 Chart Paper Loading (RD200)

Set the holes on the chart paper to the sprockets.

Omega RD200 — 5. INSTALLATION 5.1 Chart Paper Loading (RD200) diagram (page 26)
Fig. 121 · 5. INSTALLATION 5.1 Chart Paper Loading (RD200)

Don’t insert the chart paper folds reversely when inserting the chart paper into the chart paper tray, otherwise a folding failure results. 4 Check 1) Manual check Turn the thumb wheel by hand to make sure that the chart paper is feeding properly.

located on the right and left side of the internal unit. Set the guides of the chart paper cassette to the guide rails and push it until it clicks into place. (2) Return the keyboard back in its original position. 3) Chart paper feeding check (1) Turn on the power supply. (2) Press the key and check the chart

Omega RD200 — 5. INSTALLATION 5.1 Chart Paper Loading (RD200) diagram (page 26)
Fig. 122 · 5. INSTALLATION 5.1 Chart Paper Loading (RD200)

FEED

paper feeds smoothly. (3) Repeat the above procedure if the chart paper

is not fed smoothly. 4) Operation during chart end detection When the chart paper cassette is returned in the internal unit during the chart end detection, the CHART END display blinks. Press to turn it from blinking to steady lighting. Press again. The CHARD END display goes out and the chart paper is fed automatically.

Show 6 more labels / short text on this page
Thumb wheel · (Top view) · Chart paper folds · Folding section · Chart paper · -22-
Page 27 of the Omega RD200 user manual — 5. INSTALLATION 5.2 Chart Paper Loading (RD2800)
Page 27 · click to enlarge
Page 27

5. INSTALLATION 5.2 Chart Paper Loading (RD2800)

1 Chart Paper Cassette Removal 1) Open the door.

2 Chart Paper Loading 1) Open the chart paper holders. Open the chart paper guide and the stripper plate.

Omega RD200 — 5. INSTALLATION 5.2 Chart Paper Loading (RD2800) diagram (page 27)
Fig. 123 · 5. INSTALLATION 5.2 Chart Paper Loading (RD2800)
Omega RD200 — 5. INSTALLATION 5.2 Chart Paper Loading (RD2800) diagram (page 27)
Fig. 124 · 5. INSTALLATION 5.2 Chart Paper Loading (RD2800)
Omega RD200 — 5. INSTALLATION 5.2 Chart Paper Loading (RD2800) diagram (page 27)
Fig. 125 · 5. INSTALLATION 5.2 Chart Paper Loading (RD2800)

Open from here.

2. Remove the chart paper cassette. (1) Pull the grip of the chart paper cassette and
2. Prepare a chart paper. Shuffle both end of the chart paper to prevent two or more chart papers fed being together.

take the chart paper cassette slightly out of the internal unit. (Pens lift up.) (2) Pull the grip slowly to take the chart paper

cassette out of the internal unit completely.

Omega RD200 — 5. INSTALLATION 5.2 Chart Paper Loading (RD2800) diagram (page 27)
Fig. 126 · 5. INSTALLATION 5.2 Chart Paper Loading (RD2800)
Omega RD200 — 5. INSTALLATION 5.2 Chart Paper Loading (RD2800) diagram (page 27)
Fig. 127 · 5. INSTALLATION 5.2 Chart Paper Loading (RD2800)
3. Put it into the chart paper housing. The shapes of the sprocket holes on the right and left sides are different. The right side holes are an elliptical shape.
Omega RD200 — 5. INSTALLATION 5.2 Chart Paper Loading (RD2800) diagram (page 27)
Fig. 128 · 5. INSTALLATION 5.2 Chart Paper Loading (RD2800)

Chart paper cassette Guide rail

Printing side (Caution in red)

Omega RD200 — 5. INSTALLATION 5.2 Chart Paper Loading (RD2800) diagram (page 27)
Fig. 129 · 5. INSTALLATION 5.2 Chart Paper Loading (RD2800)

Be careful with the corners of the rear stripper plate Caution

The sharp corners of the rear stripper plate help to feed the chart paper smoothly. Be careful not to cut your fingers when loading or replacing the chart paper. Don’t remove the chart paper cassette during printing by a plotter pen. The cassette may hit 1st pen during the automatic pen lift-up function executed.

Don’t remove the chart paper cassette during printing.

Show 10 more labels / short text on this page
Stripper plate · Chart paper guide · Grip · Guide · Sprocket holes (circular) · Drum · Sprocket holes · (elliptical) · Chart paper housing · -23-
Page 28 of the Omega RD200 user manual — FEED
Page 28 · click to enlarge
Page 28

5. INSTALLATION 5.2 Chart Paper Loading (RD2800)

3 Chart Paper Setting 1) Draw out the chart paper approx. 50 cm and close the rear stripper plate. 2) Set the holes on chart paper over the sprockets at both ends of the drum. 3) Use the thumb wheel to advance the chart paper for 2 to 3 folds into the chart tray. 4) Close the front chart paper guide. Make sure that the holes are set over the sprockets.

4 Check 1) Manual check Turn the thumb wheel by hand to make sure that the chart paper is feeding properly.

Turning direction of the thumb wheel

Don’t turn the thumb wheel inward. The chart paper cannot be fed backward. This action may cause a chart feed failure. 2) Place the chart paper cassette in its the original position Guide rails for the chart paper cassette are located on the right and left side of the internal unit. Set the guides of the chart paper cassette to the guide rails and push it until it clicks into place. 3) Chart paper feeding check (1) Turn on the power supply. (2) Press the key and check that chart

Omega RD200 — 5. INSTALLATION 5.2 Chart Paper Loading (RD2800) diagram (page 28)
Fig. 130 · 5. INSTALLATION 5.2 Chart Paper Loading (RD2800)
Omega RD200 — 5. INSTALLATION 5.2 Chart Paper Loading (RD2800) diagram (page 28)
Fig. 131 · 5. INSTALLATION 5.2 Chart Paper Loading (RD2800)
Omega RD200 — 5. INSTALLATION 5.2 Chart Paper Loading (RD2800) diagram (page 28)
Fig. 132 · 5. INSTALLATION 5.2 Chart Paper Loading (RD2800)

○ ×

Set the holes of the chart paper on the sprockets.

FEED

paper feeds smoothly. (3) Repeat the above procedure if the chart paper

The mark for fold thread parts and the mark for fold valley parts are printed on both ends of the chart paper. Don’t insert the chart paper folds the wrong way when inserting the chart paper into the chart paper tray, otherwise a folding failure will result.

is not fed smoothly. 4) Operation during chart paper end detection When the chart paper cassette is returned in the internal unit during the chart end detection, the CHART END display blinks. Press to turn it from blinking to steady lighting. Press again. The CHARD END display goes out and the chart paper is fed automatically.

凸mark (Fold thread)

Omega RD200 — FEED diagram (page 28)
Fig. 133

凹mark 凸 mark

Chart paper No.

凹 mark (Fold valley)

Show 5 more labels / short text on this page
Thumb wheel · Chart paper folds · Time scale · Chart paper · -24-
Page 29 of the Omega RD200 user manual — REC ON/OFF
Page 29 · click to enlarge
Page 29

5. INSTALLATION 5.3 Recording Pen Loading

1 Recording Pen Types (1) There are two types of recording pens, the

plotter pens for use in digital printing and cartridge pen for use in trace printing. (2) There are four kinds of cartridge pens for 1st

Common for the RD200 and RD2800

pen to 4th pen. These are of same shape but differ in ink colors.

For RD200 For RD2800

1 st penRed 3 rd pen: Blue
2 nd penGreen 4 th pen: Brown

2 Preparation for Installation 1) To remove the chart paper cassette RD200 See Section 5.1.1. RD2800 See Section 5.2.1. 2) Open the display panel. Swing out the panel from the right edge.

  • Pen holders for RD200

4th pen 3rd pen 2nd pen 1st pen • Pen holders for RD2800

3. Set to the pen replacement mode. (1) Turn on the power supply.
Omega RD200 — 5. INSTALLATION 5.3 Recording Pen Loading diagram (page 29)
Fig. 134 · 5. INSTALLATION 5.3 Recording Pen Loading

(2) If the status RECORD ON lights, go it out by

ENTRY REC ON/OFF

pressing and then . (Pens lift

up.) (3) Press and hold down for 3 seconds.

REC ON/OFF

The pens return to original positions and then move to the center. (4) Each press of moves each pen to

4th pen 3rd pen 2nd pen 1st pen

the "zero" position. Replace each pen at its “zero” position.

Omega RD200 — REC ON/OFF diagram (page 29)
Fig. 135

1. Pen tip The pen tip is made of felt material. If an excessive force is added to it, the top of the pen tip will be crushed making clear printing or tracing impossible. 2. Pen cap Each pen is provided with a pen cap for preventing drying and protecting the pen tip. Remove and retain the pen caps before the pen installation. 3. Before stopping recording for long hours When it is expected that recording will not be executed for a day or more, remove and store the pens by attaching the pen cap to them. If the pens are left installed while no recording is executed, the ink will stain the chart paper.

Show 8 more labels / short text on this page
Plotter · pen · Cartridge · pen · Plotter pen · Plotter pen · Maintenance of pens · -25-
Page 30 of the Omega RD200 user manual — 5. INSTALLATION 5.3 Recording Pen Loading
Page 30 · click to enlarge
Page 30

5. INSTALLATION 5.3 Recording Pen Loading

3 Plotter Pen Loading (1) Before loading, remove the pen cap from

Omega RD200 — 5. INSTALLATION 5.3 Recording Pen Loading diagram (page 30)
Fig. 136 · 5. INSTALLATION 5.3 Recording Pen Loading

the plotter pen and write letters on a paper, etc. for testing. (2) Insert the plotter pen into its penholder

completely.

(Note) Incomplete insertion may result in

Omega RD200 — 5. INSTALLATION 5.3 Recording Pen Loading diagram (page 30)
Fig. 137 · 5. INSTALLATION 5.3 Recording Pen Loading

recording troubles. (3) For unloading of the plotter pen, pull it from

the penholder.

4 Cartridge Pen Loading (1) Before loading, remove the pen cap from the cartridge pen and write letters on a paper, etc. for testing. (2) Push the cartridge pen into the penholder for the same pen No. as the pen.

  • RD200
  • RD2800
Omega RD200 — 5. INSTALLATION 5.3 Recording Pen Loading diagram (page 30)
Fig. 138 · 5. INSTALLATION 5.3 Recording Pen Loading
Omega RD200 — 5. INSTALLATION 5.3 Recording Pen Loading diagram (page 30)
Fig. 139 · 5. INSTALLATION 5.3 Recording Pen Loading

Push in. 1

Omega RD200 — 5. INSTALLATION 5.3 Recording Pen Loading diagram (page 30)
Fig. 140 · 5. INSTALLATION 5.3 Recording Pen Loading
Omega RD200 — 5. INSTALLATION 5.3 Recording Pen Loading diagram (page 30)
Fig. 141 · 5. INSTALLATION 5.3 Recording Pen Loading
Omega RD200 — 5. INSTALLATION 5.3 Recording Pen Loading diagram (page 30)
Fig. 142 · 5. INSTALLATION 5.3 Recording Pen Loading

Insert first, then push in .

(3) For unloading of the cartridge pen,

  • RD200: Pull out the left side of the pen from the holder, and then take off it by disengage its convex part at the right side of the holder.
  • RD2800: Pull the pen from the holder.
Show 6 more labels / short text on this page
Plotter pen loading · Main shaft · Penholder · Plotter pen · Cartridge Pen Loading · -26-
Page 31 of the Omega RD200 user manual — CH 1 CH 2 CH 3
Page 31 · click to enlarge
Page 31

6. BASIC OPERATION 6.1 Power Supply and Operation

1 Power Supply On/Off Your recorder is not provided with a power switch. Prepare an external power switch for turning on or off its power supply. 2 Initial Operation By turning on the power supply, 1) year, month and day are displayed. After initialization, 2) year, month, day and time are printed (Printing when the power is turned on) and 3) the normal operation starts. 3 Operation 1) Operation screen (1) Measured values for all channels (CH 1 to 4: depending on the numbers of input points) are digitally displayed. (2) Measured values for all channels (CH 1 to 4: depending on the numbers of input points) are displayed by bargraphs. (3) A corresponding status lamp lights. The default is RECORD ON.

L N

Omega RD200 — L N diagram (page 31)
Fig. 143

Power and protective conductor terminals

To protective conductor of power supply facility

(250V 3A)

Example of operation screen (3-pen type)

Omega RD200 — (250V 3A) diagram (page 31)
Fig. 144

Ch 1 Ch 2 Ch 3 Chart speed

CH 1 CH 2 CH 3

(Note) The above example is for RD200. It is the same for RD2800.

2. Chart paper feeding The chart paper is fed at the programmed chart paper speed. (The default speed is 20 mm/h for RD200 and 25 mm/h for RD2800.) 3) Printing (1) Trace printing

Trace printing and fixed-time printing

Omega RD200 — CH 1 CH 2 CH 3 diagram (page 31)
Fig. 145

Measured values are traced by a cartridge pen.

CH 1: Red, CH 2: Green, CH 3: Blue, CH 4: Brown (2) Digital data printing a. Fixed-time printing

  • The following data is printed at preset time intervals.
  • Time line
  • Time
  • Year/Month/Day
  • Chart speed
  • Scale
  • Engineering unit
  • Tag b. Other printings

(Note) The above example is for RD200.

It is the same for RD2800.

  • The following printings are executed according to programming and operation.
  • Periodic data printing
  • Digital data printing
  • List 1, 2 and 3
  • Programming change mark
  • Alarm activation/reset
  • Message
  • Operation printing line and No.
  • Time axis synchronization mark
Show 8 more labels / short text on this page
Power switch installation · Overcurrent protective device · Power supply source · Power switch · Measured value · Status · Bargraph · -27-
Page 32 of the Omega RD200 user manual — 3 Chart End Detection
Page 32 · click to enlarge
Page 32

6. BASIC OPERATION 6.2 Printing ON/OFF and chart end detection

1 Printing On/Off Every time pressing then within 5 sec., the printing function switches to on or off.

Status display when printing is on

REC ON/OFF ENTRY

Omega RD200 — REC ON/OFF ENTRY diagram (page 32)
Fig. 146

RECORD ON lights.

ENTRY REC ON/OFF →

Lights when the printing is on.

RECORD OFF goes out.

No printing on/off operation executed

REC ON/OFF ENTRY →

(1) The printing on/off operation is not executed

when the status KEY LOCK lights. For its cancellation, see Section 11.13. (2) The printing on/off operation is not executed

Printing status when turning on the power supply

When turning on the power supply, the printing maintains the condition (on or off) when it was turned off last time.

when the printing is turned off by the remote contacts signal (option). See Section 13.1.

Up and down of pens

Execution by pressing key

(1) When the printing is turned off, each pen

(1) appears for about 5 sec.

lifts up automatically. Even if the power is turned off in this condition, this pen position is maintained. (2) When the chart cassette is pulled out,

when pressing key. Press

REC ON/OFF ENTRY

during this display to execute the printing on/off

operation.

each pen lifts up. If the printing has been turned off when pulling out the chart cassette, each pen has already been lifted up and it maintains its position. 2 Printing Operation The table below explains the printing operation when the printing is on or off.

(2) The printing on/off operation cannot be

executed by pressing after this display

goes out.

In such case, press and then

REC ON/OFF ENTRY

again.

3 Chart End Detection

When the chart end is detected during the printing is on, the printing is switched to off and CHART END blinks. Each pen lifts up automatically and the printing is stopped.

Chart feeding Execute Stop

For continuing the printing, refer to “Chart loading” in Section 5.1 and 5.2.

Trace printing (Cartridge pen) Execute Up where it

Up at standby position

Up and down of pens Down Up

Show 11 more labels / short text on this page
Status · Printing ON · Status · Printing OFF · Reference 1 · Reference 2 · Operation ON OFF · is · Digital data printing · (Plotter pen) Execute · -28-
Page 33 of the Omega RD200 user manual — FEED
Page 33 · click to enlarge
Page 33

6. BASIC OPERATION 6.3 Fast Feeding of Chart Paper

When pressing , chart is fed at the speed of about 0.1mm/sec. When pressing more than 1 sec, chart is fed at the speed of about 10mm/sec.This operation is used for the following purposes.

FEED

FEED

(1) Setting the time scale line of the chart paper (2) Checking of chart paper feeding operation

Fast feeding condition and operation

1. Condition ….. (1) For the fast feeding of the chart paper, the status KEY LOCK should be gone out. (2) This function cannot be executed during printing. 2. Operation ….. Digital data printing is not executed during the fast feeding of the chart paper. This function can be executed when the printing is off (RECORD ON goes out.).

Pull out the chart paper cassette slightly from your recorder. Turn the thumb wheel on the left side of the drum.

Manual chart feeding Reference 1

RD200 RD2800 Note: Even if the time scale line is set by manual chart paper feeding, a delay may occur due to the back lash of the thumb wheel (gear).

Omega RD200 — FEED diagram (page 33)
Fig. 147
Omega RD200 — FEED diagram (page 33)
Fig. 148

Time scale (1 to 24) Remarks 2

Numeric values of 1 to 24 are marked on the left side of the chart paper. These are time scales when operating the printing at a chart speed of 20 mm/h for RD200 or 25mm/h for RD2800.

Show 5 more labels / short text on this page
Thumb wheel · Thumb wheel · (Gear) · (Gear) · -29-
Page 34 of the Omega RD200 user manual — DISP DISP DISP
Page 34 · click to enlarge
Page 34

6. BASIC OPERATION 6.4 Switching Operation Screen (RD200)

For RD200, three operation screens, (1) measured value screen, (2) clock screen and (3) alarm activation screen, are available. Each time is pressed, the screen switches. Bargraphs for analog indications are displayed in any of these screens.

DISP DISP DISP

screen Clock screen Alarm activation

(Note) The alarm activation screen is only available when an alarm activates (shown by dotted lines). When

no alarm activates, this screen is skipped. 1 Measured value screen The screen differs on Model No. (number of pens). 1) 1-pen type (CH 1)

2. 2-pen type (CH 1 and 2)
Omega RD200 — DISP DISP DISP diagram (page 34)
Fig. 149

Time Ch 2 Ch 1 Chart speed

Omega RD200 — DISP DISP DISP diagram (page 34)
Fig. 150

Time Chart speed Ch 1

3. 3-pen type (CH 1 to 3) 2 Clock screen 3 Alarm activation screen This screen is only available when an alarm activates. For alarm types, see Section 11.5, item 1 .
4. 4-pen type (CH 1 to 4)
Omega RD200 — DISP DISP DISP diagram (page 34)
Fig. 151

Ch 1 Ch 2 Ch 3 Chart speed

Omega RD200 — DISP DISP DISP diagram (page 34)
Fig. 152

Ch 4 Ch 3 Ch 2 Ch 1

Screen when turning on the power supply

Omega RD200 — DISP DISP DISP diagram (page 34)
Fig. 153

Chart speed Time Day Month Year

When the power supply is turned on, the measured value screen appears.

Explanation for alarm activation

In each channel, the alarm level 1, 2, 3 and 4 are allocated to the digits from the left end. The alarm type shall be displayed only at the digit where an alarm activates.

Omega RD200 — DISP DISP DISP diagram (page 34)
Fig. 154

Alarm type at level 1

Alarm type at level 2

Alarm type at level 3

Alarm type at level 4

Show 5 more labels / short text on this page
Measured value · screen · Reference · H b · -30-
Page 35 of the Omega RD200 user manual — DISP DISP
Page 35 · click to enlarge
Page 35

6. BASIC OPERATION 6.5 Switching Operation Screen (RD2800)

For RD2800, two operation screens, (1) measured value screen and (2) alarm activation screen, are available. Each time is pressed, the screen switches. Bargraphs for analog indications are displayed in any of these screens.

DISP DISP

Measured value screen Alarm activation

(Note) The alarm activation screen is only available when an alarm activates (shown by dotted lines). When no

alarm activates, the measured value screen only appears even if is pressed.

1 Measured value screen The screen below is for 4-pen type. 1-pen type - measured value is displayed in CH 1. 2-pen type - measured values are displayed in CH 1 and CH 2. 3-pen type - measured values are displayed in CH 1 to CH 3. 2 Alarm activation screen This screen is only available when an alarm activates.

Omega RD200 — DISP DISP diagram (page 35)
Fig. 155

Ch 1 Ch 2 Ch 3 Ch 4 Time Chart speed

Omega RD200 — DISP DISP diagram (page 35)
Fig. 156
Omega RD200 — DISP DISP diagram (page 35)
Fig. 157
Omega RD200 — DISP DISP diagram (page 35)
Fig. 158

Explanation for the display

In each channel, the alarm level 1, 2, 3 and 4 are allocated to the digits from the left end. The alarm type shall be displayed only at the digit where an alarm activates. For alarm types, see Section 11.5, item 1.

Alarm type at level 1 Alarm type at level 2 Alarm type at level 3 Alarm type at level 4

screen · -31-

Page 36 of the Omega RD200 user manual — SHIFT SHIFT
Page 36 · click to enlarge
Page 36

7. PROGRAMMING 7.1 Keys and Characters

1 Keys and Their Abbreviations

Actual keys Key abbreviations used in this manual

Omega RD200 — 1 Keys and Their Abbreviations diagram (page 36)
Fig. 159 · 1 Keys and Their Abbreviations

REC ON/OFF FEED DATA PRINT DISP

SPACE COPY(=)

7 RANGE

8 ALARM

9 LIST

SET END

CLOCK

4 SCALE

5 DATA I

6 A ~ Z

SHIFT 0 CLEAR

1 CHART

2 TAG

3 UNIT

° C

° F

2 Characters Displayed in Programming Except for the leftmost digit, characters including alphabets are displayed as shown below due to the restriction with 7-segment LCD.

A B C D E F G H I J L L M N O

P Q R S T U V W X Y Z

1 2 3 4 5 6 7 8 9 0 Space Minus Hyphen Colon To

3 Key Operation

Don’t press keys with hard and sharp objects

The keys are composed of sheet switches containing electrical circuitry inside. If a key is pressed with a hard and sharp object, malfunctions due to wire disconnection or insulation breakdown may occur.

Press a single key. Press two keys simultaneously. Press two keys simultaneously for 3 sec. or longer

To program "6":

To program an alphabet letter:

To move the "Engineering" programming mode:

SHIFT 6 A - Z

+ Press ( )key first.

SHIFT 6 A - Z

6 A - Z

3 sec. or more Press ( )key first.

0.5 sec.* 3 sec. or more

SHIFT SHIFT

*Press the key for about 0.5 sec. and release it as soon as the display changes. If another screen appears by holding the key down for 3 sec. or more, press to return to the original screen and press the key for about 0.5 sec. again

Show 6 more labels / short text on this page
- (-) · % · / · 7-segment LCD · ( + ) · -32-
Page 37 of the Omega RD200 user manual — SHIFT
Page 37 · click to enlarge
Page 37

7. PROGRAMMING 7.2 Key Functions

1 Functions by Pressing Single Key

Display Stops programming and returns to the operation screen. For certain programming items, holding this key down for 2 seconds or more is required.

SPACE COPY(=)

Space (1) Deletes a numeric value or decimal point at an unnecessary digit. (2) Programs “Blank” in the programming of [engineering unit] or [tag].

Changes from programming mode (confirmation of parameters) to the programmable condition. The programming lamp lights and the cursor appears at the programmable leftmost digit.

SET END

Shift Press this key first for using the key functions being indicated on the lower case of each key.

SHIFT

CLOCK

(1) Press these keys for programming a numeric value. (2) For programming a decimal point, move the cursor to the next digit and press before entering a numeric value. (3) For deleting a decimal point, move the cursor to the next digit and press .

0 CLEAR

CLOCK

SPACE COPY(=)

9 LIST

(1) For items to program by selecting from menu, this key steps the menu forward or

backward. (2) For items to program parameters per channel in the programming mode

(confirmation of parameters), this key steps the channel No. forward or backward.

Left cursor Moves the cursor leftward.

Right cursor Moves the cursor rightward.

(1) Programming items common to channels: Stores the programmed parameter into

memory. (2) Programming items per channel: Stores temporarily the programmed parameter in the channel No. selected. Press + to store it into memory. (3) Printing on/off, data printing or list printing is executed by pressing this key.

SHIFT SET END

REC ON/OFF

on/off Selects on (enable) or off (disable) of printing. Press to go this selection active.

FEED

Feeding The chart paper is fast fed when holding this key down.

Data print Prints the measured values at the moment the key is pressed digitally on the trace printing. Press to go this printing active. 2 Functions by Pressing Two Keys Simultaneously

DATA PRINT

SHIFT

+ End Stores the programmed parameter, which is temporarily stored, into memory.

SET END

SHIFT

+ To Inserts [~()] between minimum and maximum values.

SHIFT

+ Clear Clears (Blanks) the programmed parameter displayed.

0 CLEAR

+ Alphabetic characters Each time is pressed, characters step in the order of A, B, C …..

6 A ~ Z

SHIFT

+ List printing Programmed parameters are printed as a list. Press to go this printing active. + % Programs “%” (percent) character.

SHIFT

9 LIST

SHIFT

+ / (=) Programs “/” (slash) character. This is also “=” in the subtract printing programming mode. + °C Programs “°C” (degree Centigrade) character.

SHIFT

° C

SHIFT

° F

SHIFT

+ °F Programs “°F” (degree Fahrenheit) character.

Show 22 more labels / short text on this page
Keys Names Functions · Program · - (-) · Minus · Decimal · point · Numeric · values · to · ~ · / · Up · % · Down · ℃ · Printing · Keys Names Functions · - (-) · / · % · / · -33-
Page 38 of the Omega RD200 user manual — 8 ALARM
Page 38 · click to enlarge
Page 38

7. PROGRAMMING 7.3 List of Programming Items

1 Parameters to be Programmed First

Parameters Key operations Programming

items Defaults Programming ranges Ref. page

Range No. 07 01 to 10, 21 to 56, 70 to 80 RJ enable/

disable 0 0 (disable), 1 (enable: thermocouple range only) Range/ printing

SHIFT 7 RANGE

range 0.000 to 5.000 Max. 10 digits (minimum value to maximum value). Excluding decimal point.

°C (°C calculation) °F (°F calculation) Thermocouple and resistance thermometer range only

°C/°F calculation

SHIFT

+ (3 sec. or more) - ° C

Chart speed + - 20/25 * 0001 to 0600 mm/h or 0001 to 0200 mm/min. 8.4

SHIFT 1 CHART

* RD20020 mm/h, RD2800: 25 mm/h 2 Other Programmable Parameters and Functions

Parameters Key operations Programming

items Defaults Programming ranges Ref. page

Clock + - Japanese

SHIFT

CLOCK

time 2000.01.01.00:00 to 2099.12.31.23:59 11.1

Max. 11 digits (minimum value to maximum value), excluding decimal point. Printing range reflects to the scale. Therefore no programming is normally necessary for thermocouple and resistance thermometer range.

Scale * + - 0.000 to 5.000

SHIFT

4 SCALE

  • Programming is required in [Range/Printing range] programming mode.
  • Channel to be deleted → Clear the range/printing range.
  • Channel to be recovered → Program the range/ printing range.

Skip * + - Not programmed

SHIFT

7 RANGE

Programming is required in [Range/Printing range] programming mode. [Subtract printing CH] = [Reference CH] – [subtraction CH*] * Can be replaced with [reference value (with decimal point)].

Subtract printing * + - Not programmed

SHIFT

7 RANGE

Select from CH (1 to 4), Fail* or C.End (chart end). * Recorders with alarm output (option) only.

Up to 4 alarm points can be programmed to each channel (CH 1 to 4).

Applicable to CH (1 to 4) selected in CH/output type mode. H/L (absolute value high/low limits) E/F (absolute value high/low limits with standby) U/d (rate-of-change increase/decrease limits) b/S (differential high/low limits) J/k (differential high/low limits with standby)

8 ALARM

SHIFT

Applicable to recorders with alarm output (option) Relay No. "0" to "n" can be programmed. ("0": no output), n: number of output points. For recorders without alarm output (option), the number is fixed as “0”. Alarm values

Relay No. 0

Max. 5 digits including (-) sign. Sign is not necessary when U/d, b/S or J/k is selected. Measuring

1 to 20, Programming required only when U/d is selected.

Ref. CH Not programmed

1 to 4, Programming required only when b/S and J/k are selected.

* Programming is required to each channel. For 1-pen type, programming is only for CH 1. The items without * mark are programming common to all channels. For "alarm", programming is required to each level of channels.

Show 14 more labels / short text on this page
+ · range* · Printing · CH/output · Not programmed · type · Level Not programmed · Alarm · types H · Alarm + · Not programmed · Not programmed · count · -34-
Page 39 of the Omega RD200 user manual — ° C
Page 39 · click to enlarge
Page 39

7. PROGRAMMING 7.3 List of Programming Items

Parameters Key operations Programming

items Defaults Programming ranges Ref. page

+ (3 sec. or more) - 0.1 0.1 to 9.9 (%): % programming of scale span 11.6

° C

SHIFT

Start time Not programmed 00:00 to 23:59 Periodic data printing +

5 DATA I

SHIFT

Interval time Not programmed 00H05* to 23H59 *Limited by chart speed. 11.7

Max. 5 digits (Including numeric values, alphabetic characters, %, /, °C, °F and space) Reflected by “Range No.” programmed (mV, V, °C, k)

unit * + - V

3 UNIT

SHIFT

Max. 9 digits (Including numeric values, alphabetic characters, %, /, °C, °F and space)

Tag * + - Not programmed

SHIFT 6 A - Z

1 to 5, Five kinds of messages can be programmed. Message

No. Not programmed

0 CLEAR

6 A - Z

(3 sec. or more) Message Not programmed

Max. 15 digits (Including numeric values, alphabetic characters, %, /, °C, °F and space)

non, UP bUrn, doWn bUrn For channels with voltage range, burnout is disabled (“non”), even if burnout is programmed (up or down).

SHIFT %

Burnout * + (3 sec. or more) - non

Pass code non non (not programmed). For programming, use 4 digits from 0 to 9. Passcode/

SHIFT

+ (3 sec. or more) Key lock UnLoCkEd LockEd, UnLoCkEd

0 CLEAR

9 LIST

+ (3 sec. or more) - oFF on, oFF 9.6

0 CLEAR

8 ALARM

Input filter* + (3 sec. or more)

Time constant 0 0, 0 to 10 (sec.). “0” means no time constant. 11.13

Programmed parameters can be copied to other channels. Available parameters to be copied are [Range/Printing range], [Engineering unit] and [Tag].

+ Programming in each programming mode

SHIFT SPACE COPY(=)

- -

* Programming is required to each channel. For 1-pen type, programming is only for CH 1. The items without * mark are programming common to all channels. For "alarm", programming is required to each level of channels. 3 Programming for Options

items Key operations Programming

contents Defaults Programming ranges Ref. page

"0" to "n" (n: 6, 12)* In alarm programming mode, program relay Nos. (terminal Nos.) for alarm output. * number of output points

SHIFT 8 ALARM

No. + Relay No. 0

+ (3 sec. or more)

AND /

0 CLEAR

° C

AND/OR or AND/OR programming is required for each relay No.

OR

+ (3 sec. or more) Alarm display U H (hold), U (not hold)

Relay coil E E (energize), d (not energize) Relay output U H (hold), U (not hold)

Common to all relay

° C

Nos.

Show 14 more labels / short text on this page
Alarm dead · band · Engineering · + · / · keylock · Time axis synchronization · Copy · Parameters Programming · Relay · Alarm output · Output · mode · -35-
Page 40 of the Omega RD200 user manual — SHIFT - (-)
Page 40 · click to enlarge
Page 40

7. PROGRAMMING 7.3 List of Programming Items

items Key operations Programming

contents Defaults Programming ranges Ref. page

Speed No. 1 1 to 3 Chart speed (3 speeds) +

SHIFT

1 CHART

Speed 20/25* 0001 to 0600 mm/h or 0001 to 0200 mm/min

Terminal No. 1 1 to 4

Selection from 13 items; non, printing on/off & 3 kinds of chart speed, message printing (No. 1 to 2), message printing (No. 1 to 5), operation record (A/B/C/D), digital data printing, list printing (No. 1to 3) and totalizing reset

0 CLEAR

+ (3 sec. or more) Operation non

5 DATA I

Printing No. A A, B, C, D

0 CLEAR

2 TAG

+ (3 sec. or more) Printing position 0

10 to 90%. Printing position when the contact is open. The printing position is at +5 mm when it is shorted.

Format Sd Select Ar from Sd (standard), Ar, SP, and PL.

Channel Not programmed 1 to 4

Zero for range 1 Not programmed a (Minimum value ≤ a < b)

Span for range 1 Not programmed b (a < b < c)

SHIFT

Span for range 2 Not programmed c (b < c < d)

Span for range 3 Not programmed d (c < d < e)

Span for range 4 Not programmed e (d < e < f)

Span for range 5 Not programmed f (e < f ≤ maximum value)

Format Sd Select SP from Sd (standard), Ar, SP, and PL.

Channel Not programmed 1 to 4

Printing position for broken point 1

Not programmed 0 to a (%)

Printing position for broken point 2

Not programmed a to 100 (%)

SHIFT

Scale at zero Not programmed b (Minimum value ≤ b < c)

(SP)

Not programmed c (b < c < d)

Not programmed d (c < d < e)

Scale for span Not programmed e (d

Format Sd Select PL from Sd (standard), Ar, SP, and PL.

CH for area 1 Not programmed 1 to 4

CH for area 2 Not programmed

1 to 4 (Except channels for area 1)

(PL) +

SHIFT

and 2 only for RD200

CH for area 3 Not programmed

1 to 4 (Except channels for area 1 and 2)

CH for area 4 Not programmed

1 to 4 (Except channels for area 1 to 3)

SPACE COPY(=)

15.1 Communi-

SHIFT

(3 sec. or more)

For details, refer to the separate instruction manual for “Communications Interface”.

SHIFT - (-)

(3 sec. or more)

* RD20020 mm/h, RD2800: 25 mm/h
Show 24 more labels / short text on this page
Names Programming · Terminal allocation · Remote contacts · Operation record · Automatic range shift · + · (Ar) · Record · format · Compressed/ · Expanded · + · Scale for broken · point 1 · Scale for broken · point 2 · Zone scale · Area 1 · + · Protocol · cations interface · + · Specifications · -36-
Page 41 of the Omega RD200 user manual — CLOCK
Page 41 · click to enlarge
Page 41

7. PROGRAMMING 7.4 Programming Procedures

This section explains the basic programming parameters. 1 Basic of Programming Flow Chart

Omega RD200 — 7. PROGRAMMING 7.4 Programming Procedures diagram (page 41)
Fig. 160 · 7. PROGRAMMING 7.4 Programming Procedures

Keys to be pressed depending on programming items. This action is necessary for some programming items. For parameters to be programmed for each channel,

with or . (1) Programming lamp lights. (2) The cursor appears at a

are not stored into memory. (Note) Common programmed parameters are

stored only by pressing key. (1) Memory colon starts blinking.

Programmed parameter (including temporarily stored programmed parameters) are stored. During this storing operation, measurement is interrupted. * The change mark is printed on the right side of chart paper. However the mark will not be printed completely until the chart paper advances several lines. 2 Key Operation

Use [copy] function for convenient operation. See Section 11. 14.

  • 2) How to add or delete a decimal point
  • To add the decimal point move the cursor to and press . To delete the decimal point, press . (Numeric vallue is also deleted.) 3) Clearing a displayed programmed parameter
  • The programmed parameter is cleared by pressing and simultaneously .
  • 1) Numeric keys and cursor
  • Cursor shifts rightward when pressing numeric keys ( to and ).
  • Press or for moving the cursor.

0 CLEAR 9 LIST - (-) °C °F

CLOCK

SPACE COPY(=)

SHIFT 0 CLEAR

-37-

Page 42 of the Omega RD200 user manual — SHIFT
Page 42 · click to enlarge
Page 42
Omega RD200 — 7. PROGRAMMING 7.1 Keys and Characters diagram (page 42)
Fig. 161 · 7. PROGRAMMING 7.1 Keys and Characters

An error for the programmed parameter is judged when pressing during [Storing] or [Temporarily storing] procedures. 1 Error Judgement Flow Chart

Omega RD200 — 7. PROGRAMMING 7.1 Keys and Characters diagram (page 42)
Fig. 162 · 7. PROGRAMMING 7.1 Keys and Characters
Omega RD200 — 7. PROGRAMMING 7.1 Keys and Characters diagram (page 42)
Fig. 163 · 7. PROGRAMMING 7.1 Keys and Characters

2 Type of Errors and Error Display

If a programmed parameter is in error, a long tone (approx. 0.5 sec.) sounds and no storing (or temporarily storing) is executed.

Omega RD200 — 2 Type of Errors and Error Display diagram (page 42)
Fig. 164 · 2 Type of Errors and Error Display

Programmed parameter format is in error. [1234] was programmed to [12-34] or the like, for example.

A numeric value out of programming range was programmed.

3 Remedial measures when an error occurs

If any key other than is pressed, it returns to the programmed parameter display. The cursor appears and reprogramming can be executed. Reprogram by entering correct value.

SHIFT

-38-

Page 43 of the Omega RD200 user manual — YES
Page 43 · click to enlarge
Page 43

8. BASIC PROGRAMMING 8.1 Programming Parameters Before Operation

Certain parameters are to be programmed for starting operation after turning on the power supply. 1 Turning On the Power supply

Range/ Printing range 07 (±5 V)/0.000 to 5.000 (V)

By turning on the power supply for the first time, the display and printing operation start with default parameters. Perform the basic programming to match your purpose before operation.

°C/°F selection °C calculation

Chart speed RD20020 mm/h RD2800: 25 mm/h

2 Basic Programming Parameters

Be sure to program the following three parameters before operation.

Turning on the power supply

Program the following items for each channel to be used. (1) Range No. (2) Trace printing range (3) Reference junction compensation (RJ): Enable or

disable (only for thermocouple range)

NO

Is thermocouple or resistance

YES

Select either computing unit “°C” or “°F” to measure temperature when thermocouple or resistance thermometer range is used. The default parameter is “°C”. Default speed: RD200 is set to 20 mm/h

Selecting °C/°F computation

RD2800 is set to 25 mm/h Default programming range: 0001 to 0600 mm/h or

0001 to 0200 mm/h

(1) Scale* (2) Skip (channel deletion and recovery) (3) Subtract printing (4) Alarms (absolute value, rate-of-change and differential) (5) Periodic data printing (6) Clock (7) Engineering unit (8) Tag (9) Burnout enable/disable (10) Pass code/Keylock (11) Alarm dead band (12) Message (13) Time axis synchronization

Parameters to be programmed depending on

(1) Alarm output (2) Communications interface (Note) (3) Remote contacts (4) Printing format (automatic range shift, compressed/expanded, zone record) (Note) Refer to the separate instruction manual for “Communications Interface”.

Parameters to be programmed

* When the printing range is set, that value reflects to the “Scale” setting. For the scaling with voltage input, read the Section 11.2.

Show 9 more labels / short text on this page
Default parameters · Programming range/printing · range* · thermometer range used? · Programming chart speed · Basic programming end · functions · depending on options · -39-
Page 44 of the Omega RD200 user manual — SHIFT
Page 44 · click to enlarge
Page 44
Omega RD200 — 7. PROGRAMMING 7.1 Keys and Characters diagram (page 44)
Fig. 165 · 7. PROGRAMMING 7.1 Keys and Characters

Program the following parameters before operation. Press and simultaneously at the operation screen to display the “Range/Printing range” programming screen. Programming of the parameter is required in every channel. 1 Programming Mode and Parameters

1. Programming mode

“R” stands for “Range”.

Range No.

Printing range (Max. 10 digits by left-justify) *

(within 5 digits) to Maximum value

(within 5 digits) Value to print

* Unused digits should be filled with spaces.

0% + Value to print

SHIFT

2. Parameters Program the following three parameters for every channel.

Parameters (1) Range No. (2) RJ selection (3) Printing range

Purpose Selection of input types and measuring ranges

Selection of reference junction compensation (RJ) enabled or disabled

Specifying of the input range to be printed on the chart paper.

Minimum value to maximum value (within 5 digits) This range can be programmed irrespective of the measuring range being specified in the range No. selected. (Note) (Note) Input exceeding the measuring range or printing range is printed as an over-range. 2 Cautions on Programming and Reference

Range No. (2-digit numeric value) For a range No. table, see item 4 .

0. Disable (external compensation) 1: Enable (internal compensation) Program to “0” for all inputs other than thermocouple input.

Read the following cautions and reference items. The programming flow charts are described on the next page. 1) Minimum printing range Read the “minimum printing range” on Section 22.1 (Input Specifications). A programming error occurs if the minimum value and the maximum value are set to the same figure. 2) Position of decimal point The printing range reflects to the scale value. The decimal point position on actual display/printing is fixed by the range No. selected. For changing it, refer to the scale programming.

Range No. (Measuring range) Printing range Scale

07 (-5.000 to 5.000 V) 0 to 5 0 to 5 0.000 to 5.000

21 (-200.0 to 300.0℃) 0 to 200 0 to 200 0.0 to 200.0

23 (-200 to 1370℃) 0.0 to 800.0 0.0 to 800.0 0 to 800

3. When “Range/Printing range” is changed, the last scale value becomes ineffective and is replaced with the

new printing range.

4. Channel for subtract printing Different range programming procedure is required. Read Section 11.4 (Programming Subtract Printing).
5. Programming a same range to other channels The [copy] function is useful. Read Section 11.14.
6. Scaling Printing range reflects to the scale value. For scale programming to the channel with voltage input, read Section 11.2.
Show 15 more labels / short text on this page
Character display · Programming lamp · Memory colon · Character display · Channel · RJ selection · Minimum value · - (-) · 100% · Programming · values · Actual position · of decimal · point · -40-
Page 45 of the Omega RD200 user manual — YES
Page 45 · click to enlarge
Page 45

8. BASIC PROGRAMMING 8.2 Range/Printing Range

3 Programming Flow Chart

Range No. for channel 3: 22, RJ: Enable, Printing range: -50 to 450°C

Channel 1 range.

  • Other channels can be checked with or .
  • Cursor and programming lamp light.
  • Cursor moves rightward when pressing a numeric key.

7 RANGE

SHIFT

SET

END

Cursor moves rightward.

[To be ready for programming]

Cursor moves leftward.

Press or , and select [3], then press . Press and then press again (range No. 22). See the next page for the range No. table. RJ (Reference Junction compensation):

2 TAG 2 TAG

[Selecting range No.]

2 TAG 2 TAG

Range No.

1 CHART

0 → Disable* 1 → Enable *For all inputs other than thermocouple input, program to “0”.

  • Adding/deleting decimal point: Move the cursor rightward and press or .
  • Programming “to”: Press and simultaneously.
  • Decimal point is not included in the number of programming digits.

Printing range *

・ CLOCK

SPACE COPY(=)

YES

SHIFT - (-)

NO

SHIFT

An error occurs if a value exceeds the programming range or unnecessary space is entered. Display moves to the next channel for programming. Store the [temporarily stored] programmed values into memory.

YES

NO

SHIFT SET

END

Memory colon blinks and the programming change mark is printed.

* Unused digits should be filled with spaces.

Show 23 more labels / short text on this page
+ · / % · [Check] · Channel · ℃ · → · % · [Selecting channel] · % / · / · Channel · → · [Programming RJ selection] · RJ selection · Function keys · [Programming printing range] · Error message appears? · Keys other than · Temporarily stored · Programming other channels? · + · Stored · -41-
Page 46 of the Omega RD200 user manual — 8. BASIC PROGRAMMING 8.2 Range/Printing Range 4 Range No. Tables
Page 46 · click to enlarge
Page 46

8. BASIC PROGRAMMING 8.2 Range/Printing Range 4 Range No. Tables

1. Voltage input № Input type Measuring range Remarks 01 -13.80 to 13.80 mV 02 -27.60 to 27.60 mV 03 -69.00 to 69.00 mV 04 -200.0 to 200.0 mV 05
Omega RD200 — 8. BASIC PROGRAMMING 8.2 Range/Printing Range 4 Range No. Tables diagram (page 46)
Fig. 166 · 8. BASIC PROGRAMMING 8.2 Range/Printing Range 4 Rang…
Omega RD200 — 8. BASIC PROGRAMMING 8.2 Range/Printing Range 4 Range No. Tables diagram (page 46)
Fig. 167 · 8. BASIC PROGRAMMING 8.2 Range/Printing Range 4 Rang…

Measuring range No. Input type °C °F 40 W-WRe26 0 to 2315 32 to 4200 41 WRe5-WRe26 0 to 2315 32 to 4200

-500.0 to 500.0 mV 06 -2.000 to 2.000 V 07 -5.000 to 5.000 V 08 -10.00 to 10.00 V 09 -20.00 to 20.00 V 10

44 -50.0 to 290.0 32.0 to 550.0 45 -50.0 to 600.0 32 to 1110 46 NiMo-Ni

50 to 1310 32 to 2350 47 CR-AuFe 0.0 to 280.0K 0.0 to 300.0K 48 0.0 to 350.0 -140.0 to 660.0 49 0.0 to 650.0 -140 to 1200 50

2. Thermocouple input (*1) Measuring range No. Input

Ⅱ 0 to 1390 -140 to 2530 51 -200.0 to 250.0 -320.0 to 480.0 52 -200.0 to 500.0 -320.0 to 930.0 53 U

type °C °F 21 -200.0 to 300.0 -320.0 to 570.0 22 -200.0 to 600.0 -320 to 1110 23 K

-200.0 to 600.0 -320 to 1110 54 -200.0 to 250.0 -320.0 to 480.0 55 -200.0 to 500.0 -320.0 to 930.0 56 L

-200 to 1370 -320 to 2490 24 -200.0 to 200.0 -320.0 to 390.0 25 -200.0 to 350.0 -320.0 to 660.0 26 E

-200 to 900 -320 to 1650 3) Resistance thermometer input (*2) Measuring range No. Input

-200 to 900 -320 to 1650 27 -200.0 to 250.0 -320.0 to 480.0 28 -200.0 to 500.0 -320.0 to 930.0 29 J

Omega RD200 — 8. BASIC PROGRAMMING 8.2 Range/Printing Range 4 Range No. Tables diagram (page 46)
Fig. 168 · 8. BASIC PROGRAMMING 8.2 Range/Printing Range 4 Rang…

type °C °F 70 -140.0 to 150.0 -220.0 to 300.0 71 -200.0 to 300.0 -320.0 to 550.0 72

-200 to 1200 -320 to 2190 30 -200.0 to 250.0 -320.0 to 480.0 31 T -200.0 to 400.0 -320.0 to 750.0 32 0 to 1200 32 to 2190 33 R 0 to 1760 32 to 3200 34 0 to 1300 32 to 2370 35 S 0 to 1760 32 to 3200 36 B 0 to 1820 32 to 3300 37 -200.0 to 400.0 -320.0 to 750.0 38 -200.0 to 750.0 -320 to 1380 39 N

Pt100 (JIS’97) -200.0 to 850.0 -320 to 1560 73 -140.0 to 150.0 -220.0 to 300.0 74 -200.0 to 300.0 -320.0 to 550.0 75

JIS’89 -200.0 to 649.0 -320 to 1200 76 -140.0 to 150.0 -220.0 to 300.0 77 -200.0 to 300.0 -320.0 to 550.0 78 JPt100

-200.0 to 649.0 -320 to 1200 79 Pt50 -200.0 to 649.0 -320 to 1200 80 Pt-Co 4.0 to 374.0K 4.0 to 374.0K *2: (1) No.70 to 72: IEC751(1995), JIS C 1604-1997 (2) No.73 to 75: IEC751(1983), JIS C 1604-1989

-200 to 1300 -320 to 2370 *1: (1) No.21 to 39: IEC584, JIS C 1602-1995 (2) No.40 to 50: ASTM (3) No.51 to 56: DIN43710

JIS C 1606-1989 (3) No.76 to 78: JIS C 1604-1981

JIS C 1606-1989 (4) No.79JIS C 1604-1981
Omega RD200 — 8. BASIC PROGRAMMING 8.2 Range/Printing Range 4 Range No. Tables diagram (page 46)
Fig. 169 · 8. BASIC PROGRAMMING 8.2 Range/Printing Range 4 Rang…
Decimal point positionEven if the decimal point position is programmed in the printing range, the actual

display/printing position is shown as described in the table above depending on the range No. For changing the decimal point position, see Section 11.2 “Scale programming”. Engineering units: Engineering unit is shown as described in the table above depending on range No.

(Temperature unit is fixed to “°C” except No. 47 and 80). For changing the engineering unit, see Section 11.8 “Unit programming”.

-42-

Page 47 of the Omega RD200 user manual — END
Page 47 · click to enlarge
Page 47

8. BASIC PROGRAMMING 8.3 ° C/ ° F Computation Selection

This programming is to select a “°C” or “°F” computation for a temperature range (thermocouple or resistance thermometer). The default setting is “°C”. Press and simultaneously 3 seconds or more in the operation screen to display the “°C/°F computation” selection screen. Select “°F” if necessary. 1 Programming Mode

SHIFT

2 Programming Flow Chart

From “°C” computation to “°F” computation

°C= (°F - 32), °F= x °C + 32 The selected computation unit can be checked. The unit in the left flow is “°C”. Cursor and programming lamp light. Each time or is pressed, 「 」or「 」is displayed alternately. The selected computation unit is stored into memory.

“°C” ↔ “°F” conversion formula

5 9 9 5

SHIFT

3 sec. or more

SET

[To be ready for programming]

END

[Selecting “°F”]

Memory colon blinks and the programming change mark is printed.

Relation with other programming items Remarks

1) In case of temperature unit “K” The temperature unit for the range No.47 (AuFe-CR) and 80 (Pt-Co) is “K”. It does not affect this programming. 2) Influence onto other programmed parameters The programmed parameters for “Range/Printing range”, “Scale” and “Alarm setpoint” are not changed automatically. If the programmed parameters cannot be used for the selected computation unit, reprogram them again. Even if “°C” or “°F” is programmed at the “Engineering unit programming”, it does not change the computation.

Show 14 more labels / short text on this page
Character display · Programming lamp · Cursor Character · Computation unit (Fixed) · display · Memory colon · ( + ) · [Check] · Computation unit · or · / % · % / · Stored · -43-
Page 48 of the Omega RD200 user manual — YES
Page 48 · click to enlarge
Page 48

8. BASIC PROGRAMMING 8.4 Chart Speed Programming

  • Program the chart speed before operation.
  • Press and simultaneously in the operation screen to display the [Chart speed] programming screen.
  • For the remote contacts (option), read Section13.3.
  • 0000mm/h can be set for RD2800. When 0000mm/h is set, chart is fed at the speed of 12.5mm/h.

SHIFT 1 CHART

0001 to 0600 (mm/h) or 0001 to 0200 (mm/h)

1 Programming Mode

2 Programming Flow Chart

Speed unit (H or M)

Cursor Chart speed (4 digits programming) (Fixed)

( H/M )

From 20 mm/h to 50 mm/h

Changing the speed unit (h/m)

Move the cursor to the character display (a digit on the left) then press or to switch “M” and “H” alternatively.

SHIFT 1 CHART

  • Programmed chart speed
  • Cursor and programming lamp light.
  • Cursor moves rightward when pressing a numeric key.

SET

[To be ready for programming]

END

Cursor moves rightward.

Cursor moves leftward.

0 CLEAR 5 DATA I

0 CLEAR 0 CLEAR

Program the chart speed within 0001 to 0600 mm/h or 0001 to 0200 mm/min.

An error occurs if a value exceeds the programming range or any space or decimal point is entered. The new programmed value is stored into memory.

YES

SHIFT

NO

Memory colon blinks and the programming change mark is printed.

If 150 mm/h or more is programmed for the chart speed

No printing, except time line, data printing and programming change mark, is executed.

Influence to periodic data printing

When changing the chart speed, the programmed parameters for [periodic data printing] (Section 11.7) is cleared. (Default: Not programmed)

Show 16 more labels / short text on this page
Programming range · Character display · Programming lamp · Speed unit · Memory colon · % / · + · [Check] · Speed unit · Chart speed · ℃ · [Programming chart speed] · Error message appears? · Keys other than · Stored · -44-
Page 49 of the Omega RD200 user manual — ⎯
Page 49 · click to enlarge
Page 49

9. PRINTINGS 9.1 Printing Types and the Details

Printing comprises trace printing and digital printing. The fixed-time printing and trace printing are executed even if any operation is not programmed. 1 Printing Types and the Details

Printing type Printing details Remarks

Trend printing is executed for each pen (channel)

Trace printing (Cartridge pen)

1st pen 2nd pen 3rd pen 4th pen

⎯

Red Green Blue Brown

The following details are printed on the fixed time. (1) Year/Month/Day (2) Time (3) Time line (4) Chart speed (5) Scale, Engineering unit, Tag 10-1

Measured values are printed on the trace printing with programmed intervals. The printing starts from the programmed start time line. The printing mode is different depending on chart speed.

Measured values are printed on the trace printing. The printing mode is different depending on chart speed. 9-2

Major parameters (Range/Printing range, Scale, Subtract printing, Periodic data printing) are printed as a table. 9-3

The parameters (such as time axis synchronizing, message, items relating to options) other than those involved in List 1 are printed as a table. 9-3

Digital Data Printing (Plotter pen)

List 3 printing All parameters (List 1 + List 2) are printed as a table. 9-3

Message is printed on the right side of the chart paper by selecting the programmed message No. 9-4

Alarm details (time, alarm points, alarm type) are printed on the right side of the chart paper when an alarm activates or an alarm is reset. 10-4

Time axis sync. mark

(1) When the time axis synchronization (POC) is switched on or off, its time, mark

(*), and ON or OFF are printed on the right side of the chart paper. (2) When it is ON, a mark (*) is printed to the right of the time print of fixed-time

printing.

Year/Month/Day and time are printed on the left side of the chart paper when the power is turned on. 6-1

For the remote contacts (option), it is required to program the "Operation printing position" and select one of Act 1 to 4 in the "Remote contacts" programming. (1) Printing line: When a contact signal (on) is applied to the allocated terminal,

and Operation printing No.

the line is printed by 5 mm to the right of the programmed printing line position. (2) Operation printing No.: Operation printing No. (A to D) is printed at constant

intervals to the left of the programmed printing line position.

When hardware check is executed by selecting the printer (plotter pen) characters are printed for checking purpose. 19-2

Show 18 more labels / short text on this page
Fixed-time · printing · Periodic data printing · Digital data · printing · List 1 printing · List 2 printing · Message · printing · Alarm printing · 9-6 10-1 · printing · Power-on · printing · Operation printing line · Printer · check printing · -45-
Page 50 of the Omega RD200 user manual — 2 Printing Examples
Page 50 · click to enlarge
Page 50

9. PRINTINGS 9.1 Printing Types and the Details

2 Printing Examples

1. RD200 (Printing width: 100 mm)
Omega RD200 — 2 Printing Examples diagram (page 50)
Fig. 170 · 2 Printing Examples
2. RD2800 (printing width: 180 mm)
Omega RD200 — 2 Printing Examples diagram (page 50)
Fig. 171 · 2 Printing Examples
Omega RD200 — 2 Printing Examples diagram (page 50)
Fig. 172 · 2 Printing Examples
Omega RD200 — 2 Printing Examples diagram (page 50)
Fig. 173 · 2 Printing Examples

4-pen type · 4-pen type · -46-

Page 51 of the Omega RD200 user manual — REC ON/OFF ENTRY
Page 51 · click to enlarge
Page 51

9. PRINTINGS 9.2 Digital Data Printing

The measured value at the moment pressed is digitally printed with the plotter pen.

DATA PRINT

(1) Press and then press . Digital printing of

DATA PRINT ENTRY

measured value at the moment pressed starts.

DATA PRINT DATA PRINT

DATA PRINT ENTRY

(Note) When is pressed, appears

for about 5 sec. Press while this message appears.

  • Printing mode is different depending on the chart speed.

Digital data printing starts.

Chart speed Printing mode

YES*

0 to 499mm/h Analog printing is continued without interruption. 500 mm/h or more

Analog printing is interrupted and data printing is started. (2) For interrupting the printing, press and then

REC ON/OFF ENTRY

NO

REC ON/OFF

Digital data printing ends. (Approx. 5 min. at 20 mm/h)

press . Your recorder returns to the normal operation but the printing operation is turned off. For turning on the printing operation, press and then press again. (3) When the digital data printing completes, your recorder returns

* With this procedure, the printing operation turns off. Press and again to turn it on.

REC ON/OFF ENTRY

to the normal operation.

REC ON/OFF ENTRY Digital data printing cannot be executed in the following conditions.

(1) RECORD ON status lamp is not lit. (2) KEY LOCK status lamp lights.

Example of Digital data printing (4-pen type)

Omega RD200 — REC ON/OFF ENTRY diagram (page 51)
Fig. 174
Omega RD200 — REC ON/OFF ENTRY diagram (page 51)
Fig. 175
Omega RD200 — REC ON/OFF ENTRY diagram (page 51)
Fig. 176
Omega RD200 — REC ON/OFF ENTRY diagram (page 51)
Fig. 177

(Note) The above examples are for RD200. They are the same for RD2800.

Show 4 more labels / short text on this page
→ (Note) · Interrupt? · → · -47-
Page 52 of the Omega RD200 user manual — REC ON/OFF
Page 52 · click to enlarge
Page 52

9. PRINTINGS 9.3 List Printing

The details of programming such as "Range/Printing range" and "Scale" can be printed as a list with the plotter pen. As this includes large amount of data and takes time, it enables to divide the data into two and select printing of "List 1" or "List 2".

List 1: Major parameters …. “Range/Printing range”, “Chart speed”, “Printing format”, etc. List 2: Parameters other than the above …. “Message”, “Time axis synchronization.” and “option-related-items” List 3: All parameters …. “List 1” + “List 2”

(1) Press and simultaneously, and then

1 List 1 Printing

SHIFT 9 LIST

press . (2) The selection screen from [List 1] to [List 3] appears.

SHIFT 9 LIST

Move the cursor to “1”. (3) When is pressed,

° C ° F

[List No. selection]

appears for about 5 sec. For execution, go to procedure (4). For no execution, press . (4) Press to start “List 1” printing. The cartridge

(3) (4) → ENTRY ENTRY

pen moves to the end of the left side and the chart paper forwards slightly. (5) For interrupting the printing, press and

List 1 printing starts.

REC ON/OFF ENTRY

simultaneously. Your recorder returns to the normal operation but the printing operation is turned off. For turning on the printing operation, press and then press again. (6) When the list printing completes, your recorder

YES*

REC ON/OFF ENTRY

→ REC ON/OFF ENTRY

NO

returns to the normal operation.

List 1 printing ends. (Approx. 30 min. for printing)

List 1 printing details

* With this procedure, the printing operation turns off. Press and then press again to turn it on.

(1) Year/month/day, Time (2) Chart speed (Note) (3) Range/Printing range (4) Scale (5) Subtract printing (6) Periodic data printing (7) Printing format (8) Alarm (Note) When the remote contacts (option), which can

REC ON/OFF

program 3 speeds (No.1 to No.3), is added, the presently selected speed is printed.

List 1 printing cannot be executed in the following conditions.

Example of the List 1 printing

(1) RECORD ON status lamp is not lit. (2) KEY LOCK status lamp lights. Chart speed does not affect to this operation. Chart paper forwards at a fixed speed.

Omega RD200 — REC ON/OFF diagram (page 52)
Fig. 178

(1) Trace printing with a cartridge pen is

Operation during printing Reference

interrupted. Other functions are continued without being interrupted. (2) Programming any parameters

(Range/Printing range, Scale, Alarm, etc.) cannot be executed during printing. Program them after the printing ends.

(Note) The above example is for RD200. It is the same for

RD2800.

Show 8 more labels / short text on this page
+ · (1) · or · (2) · Interrupt? · (5) · (6) · -48-
Page 53 of the Omega RD200 user manual — RD2800. 3 List 3 Printing
Page 53 · click to enlarge
Page 53

9. PRINTING 9.3 List Printing

2 List 2 Printing

(1) Press and simultaneously, and

SHIFT 9 LIST

then press . (2) The selection screen from [List 1] to [List 3]

SHIFT 9 LIST

appears. Move the cursor to “2”. (3) When is pressed,

[List No. selection]

appears for about 5 sec. For execution, go to procedure (4). For no execution, press . (4) Press to start “List 2” printing. The

(3) (4) → ENTRY ENTRY

cartridge pen moves to the end of the left side and the chart paper forwards slightly. (5) For interrupting the printing, press and

List 2 printing starts.

REC ON/OFF ENTRY

simultaneously. Your recorder returns to the normal operation but the printing operation is turned off. For turning on the printing operation, press and then press again. (6) When the list printing completes, your recorder

YES*

REC ON/OFF ENTRY

→ REC ON/OFF ENTRY

NO

List 2 printing ends. (Approx. 30 min. for printing)

returns to the normal operation.

List 2 printing details

* With this procedure, the printing operation turns off. Press and then press again to turn it on.

(1) Year/month/day, Time (2) Chart speed (Note) (3) Alarm outputs (4) Time axis sync (5) Message (6) Operation record (7) Remote contacts (8) Mathematics (Note) For the remote contacts (option), which can

REC ON/OFF ENTRY

program 3 speeds (No.1 to No.3), is added, all programmed 3 speeds are printed.

List 2 printing cannot be executed in the following conditions.

(1) RECORD ON status lamp is not lit. (2) KEY LOCK status lamp lights. Chart speed does not affect this operation. Chart paper forwards at a fixed speed.

Example of the List 2 printing

Omega RD200 — REC ON/OFF ENTRY diagram (page 53)
Fig. 179

Trace printing with a cartridge pen is not interrupted. Other functions are continued without being interrupted.

Operation during printing Reference

(Note) The above example is for RD200. It is the same for

RD2800. 3 List 3 Printing

Move the cursor to “3” at the [List No. selection] and execute the same procedure as item 2. All of the programmed parameters are printed.

Show 9 more labels / short text on this page
+ · (1) · or · ℃ · (2) · Interrupt? · (5) · (6) · -49-
Page 54 of the Omega RD200 user manual — REC ON/OFF ENTRY
Page 54 · click to enlarge
Page 54

9. PRINTING 9.4 Message Printing

For printing of a pre-programmed message, select the message No. and press .

(1) Press and simultaneously for 3

0 CLEAR

6 A - Z

0 CLEAR 6 A - Z

seconds or more.

3 sec. or more

(2) Select the message No. (1 to 5) to be printed with

[No. selection]

or .

(Note) Refer to Section 11.10 for programming of

Message No.

message.

(3) Press to start “Message” printing with a

Message printing starts.

plotter pen on the right side of the chart paper.

(Note) The “message” will not be printed completely

until the chart paper forwards several lines.

YES*

(4) For interrupting the printing, press and then

REC ON/OFF

REC ON/OFF

press . The printing operation is turned off

NO

and your recorder returns to the normal operation. For turning on the printing operation, press

REC ON/OFF ENTRY

Message printing ends.

and simultaneously again.

* With this procedure, the printing operation turns off. Press and then press again to turn it on.

(5) When the message printing completes, your

recorder returns to the normal operation.

REC ON/OFF ENTRY

Example of message printing

Omega RD200 — REC ON/OFF ENTRY diagram (page 54)
Fig. 180

Message printing cannot be executed in the following conditions.

(1) RECORD ON status lamp is not lit. (2) KEY LOCK status lamp lights.

(3) For the chart speed programmed at 150

mm/h or faster When the remote contacts (option) is added, message printing can also be executed by a contact signal. However, this requires the programming of the "terminal allocation". See Section 13.2 for details.

Execution with external contact signal Reference

(Note) The above example is for RD200. It is the same for

RD2800.

Show 11 more labels / short text on this page
( + ) · (1) · % / · or · / % · (2) · Interrupt? · (4) · → · (5) · -50-
Page 55 of the Omega RD200 user manual — 9. PRINTING 9.5 Printing Format Selection
Page 55 · click to enlarge
Page 55

9. PRINTING 9.5 Printing Format Selection

This selection is only applicable to the printing format function (option). Four types of analog printing format are available. 1 Types of Printing Format

Omega RD200 — 9. PRINTING 9.5 Printing Format Selection diagram (page 55)
Fig. 181 · 9. PRINTING 9.5 Printing Format Selection

Before the selection, the programming of the desired format is required. See Section 14. The default is the standard format ().

Decimal point appears in the current format. (1) : Standard (2) : Automatic range shift (3) : Compressed/expanded (4) : Zone 1. To store the standard format; Press key. 2. To store the format other than the standard: Press and then press and simultaneously. The selected format is stored into memory. The trace printing is executed in the stored format. (Programming change mark is printed.)

-51-

Page 56 of the Omega RD200 user manual — types. 2 Flow Chart
Page 56 · click to enlarge
Page 56

9. PRINTING 9.6 Time Axis Synchronization (POC)

1 Time Axis Synchronization (POC)

Omega RD200 — 1 Time Axis Synchronization (POC) diagram (page 56)
Fig. 182 · 1 Time Axis Synchronization (POC)

(1) This function is applicable to 2-pen, 3-pen and 4-pen

types. In the standard printings, data at the same moment are printed on the chart paper with time axis gaps due to the difference in pen positions between 1st pen and 2nd to 4th pens. (When POC = OFF) (2) The time axis synchronization functions to store the data

of 2nd to 4th pens for the gaps and to print the stored data after the chart paper is fed to the printing start point of the 1st pen. (When POC = ON) (3) The time axis synchronization is not available in 1-pen

types. 2 Flow Chart

Omega RD200 — types. 2 Flow Chart diagram (page 56)
Fig. 183 · types. 2 Flow Chart

for Selecting Time Axis Synchronization From time axis synchronization Off to On

Omega RD200 — types. 2 Flow Chart diagram (page 56)
Fig. 184 · types. 2 Flow Chart

data for the gaps to 1st pen are stored in memory. G2: The data for the gaps to 1st pen,

which have been stored in memory in the printing off, are printed. G: The data for the gaps is erased. An

Selected time axis synchronization (on or off) is stored into memory. (Programming change mark is printed.) A marking as shown below is printed when the time axis synchronization is switched to on or off.

-52-

Page 57 of the Omega RD200 user manual — 1:TIC1 0.0/200.0/400.0/500.0 ℃
Page 57 · click to enlarge
Page 57

10. OPERATIONS 10.1 Fixed-Time Printing and Intervals

After your recorder is turned on, parameters including time, chart speed and scale are printed at specified intervals. The place of printing is in principle on the left side of the chart paper.

Printing items Printing intervals, etc.

1. Year/Month/Day, Time When power or printing is switched on 2000.10.03.14:25 * ( * : Only when time axis synchronization is on) 2. Year/Month/Day 24-hour period (Printing every 00 hour 00 minute) 2000.10.04
3. Chart speed Approx. 84 mm intervals 20 mm/h

Approx. 42 mm intervals, in order of channel No. When the printing format (option) is added, the printing contents vary depending on the selected format as shown in the following examples. When the printing format is not added (standard), the printing is as shown in (1).

(1) Standard (Sd) (2) Automatic range shift (Ar)

1. TIC1 0.0/500.0 ℃ ( (Note) In 5 ranges (R1 to R5), the

Channel Tag Engineering unit Scale

4. Scale 5. Engineering unit 6. Tag (Note) (Note) For only channels

1:TIC1 0.0/500.0 ℃

range when the printing is being executed is printed as the scale.

(3) Compressed/expanded (SP) (4) Zone record (PL)

+ + … ∗

+ + …. ∗

1:TIC1 0.0/200.0/400.0/500.0 ℃

1:TIC1 0.0/500.0℃ *: The “+” marking identifying zones is printed at its boundary.

2nd break point Span

* The “+” mark is printed at the positions of break points 1 and 2.

Printing of “time line” and “time” is interlocked with chart speed and is executed at the following intervals. The start point of intervals is 00:00 hours.

Chart speed (mm/h) Time line Time

1 to 9 6 hours 12:00 only

10 to 15 2 hours 4 hours

16 to 30 1 hour 2 hours

7. Time line 8. Time

31 to 60 1 hour 1 hour

61 to 119 30 minutes 1 hour

120 to 149 30 minutes 30 minutes

150 or faster 30 minutes No printing

c Time line d Time e POC mark (Note: Printed only when the time axis synchronization is on)

<Example> _ 1030 *

(Note) See Section 9.6 for details.

When the chart speed is 150 mm/h or faster Reference

No printing is executed except time line, data and programming change mark.

Show 5 more labels / short text on this page
being programmed · Zero 1st break · point · c d e · -53-
Page 58 of the Omega RD200 user manual — 1 Over-range input
Page 58 · click to enlarge
Page 58

10. OPERATIONS 10.2 Operations at Abnormal Inputs

1 Over-range input

Omega RD200 — 1 Over-range input diagram (page 58)
Fig. 185 · 1 Over-range input
  • The following table shows displays and printings for input exceeding a printing range or a measuring range.
  • Measuring range: Determined by the range No. programmed in “Range/Printing range”. See Section 8.2.
  • Printing range: Trace printing range programmed in “Range/Printing range”.
Omega RD200 — 1 Over-range input diagram (page 58)
Fig. 186 · 1 Over-range input

Display Printing 区分 Input Digital Bar graph Digital Analog

* For the digital display and printing, the measured values can be displayed and printed up to about ±10% of the span even when the values are outside the measuring range. 2 Input Disconnection

The display and printing, when the input is disconnected, differ depending on the “Burnout enable/disable” programming. However, for channels of a voltage (mV or V) range, the burnout function is “disable (non)” even when “Burnout Enable/Disable” is programmed. Program “Burnout enable/disable” to each channel. See Section 11.11.

Omega RD200 — 1 Over-range input diagram (page 58)
Fig. 187 · 1 Over-range input

Display Printing Programming burnout Digital Bar graph Digital Analog

-54-

Page 59 of the Omega RD200 user manual — 1 Display of Alarm Activation
Page 59 · click to enlarge
Page 59

10. OPERATIONS 10.3 Alarm Display and Printing

1 Display of Alarm Activation

The alarm activation information can be checked in the operation screen.

1. Measured value display screen

<Ex.> Alarm activated at CH 1 in 3-pen type Display in case alarm activated

(1) ALARM status lamp lights. (2) The measured value of the channel in

Omega RD200 — 1 Display of Alarm Activation diagram (page 59)
Fig. 188 · 1 Display of Alarm Activation

alarm blinks. (To cancel blinking) Press . The measured value is displayed steadily. (1) The character display shows “”. (2) The alarm type at the level, of which

(Note) The above example is for RD200. It is the same for

RD2800. 2) Alarm display screen

<Ex.> Alarm activated at CH 1 in 3-pen type

Display in case of alarm activated

Omega RD200 — 1 Display of Alarm Activation diagram (page 59)
Fig. 189 · 1 Display of Alarm Activation

the alarm is activated, is displayed for each channel. (3) ALARM status lamp lights. (Note) If no alarm activates, the alarm

display screen is not displayed even when is pressed. 2 Alarm Activation/Reset Printing

(Note) The above example is for RD200. It is the same for

RD2800.

Alarm activation and reset are printed on the right side of the chart paper.

1. Printing format when an alarm activates 1 3 : 5 9 1 H 1
2. Printing format when an alarm is reset 1 4 : 0 5 1 - 1

Printing example of alarm activation

Printing example of alarm reset

Omega RD200 — 1 Display of Alarm Activation diagram (page 59)
Fig. 190 · 1 Display of Alarm Activation
Omega RD200 — 1 Display of Alarm Activation diagram (page 59)
Fig. 191 · 1 Display of Alarm Activation
Show 14 more labels / short text on this page
(2) · (1) · (2) · (1) · (3) · Time · Time · Channel · Channel · Type · Hyphen · Level · Level · -55-
Page 60 of the Omega RD200 user manual — NO
Page 60 · click to enlarge
Page 60

11. OTHER PROGRAMMING 11.1 Time

For programming date and time, press and simultaneously to display the “Clock” programming screen. The default is Japanese time. 1 Programming Mode

SHIFT ・ CLOCK

2 Programming Flow Chart

Month Year Day Hour Minute

Clock data is maintained with a lithium battery. The clock keeps on working during the time that the power supply is off or if the power supply is interrupted. The service life of the battery is for eight years under the condition of eight hours operation in a day. Time display stops during programming. Press to start the display again. The time is programmed with a 24- hour display, which means that the programming range is 00:00 to 23:59.

From 00:00 hours on January 1st, 2000 to 15:40

hours on October 19th, 2000

SHIFT ・ CLOCK

SET END

[To be ready for programming]

[Programming Year, Month/Day, Time]

YES

Any key other than SHIFT

NO

Show 18 more labels / short text on this page
Character · display · Time · Clock circuit · + · [Check] · Time display during · programming Reference 1 · Numeric · keys · Year · Month Day Minute · Hour · Reference 2 · Programming time · Error message appears? · Stored · -56-
Page 61 of the Omega RD200 user manual — 3 Example of Decimal Point Programming
Page 61 · click to enlarge
Page 61

11. OTHER PROGRAMMING 11.2 Scale

“Scale” programming is necessary to display a voltage input from a converter, etc. with an actual scale. However, the scale is programmed with the same characteristic scale (not linearized scale) as the voltage input. This programming is also necessary when the decimal point position is changed in thermocouple or resistance thermometer input channels. Press and simultaneously in the operation screen to display the “scale” programming. 1 Programming Scale

SHIFT 4 SCALE

Concept of scale programming

Range No.03 (±69 mV) Printing range: Changing the programming from “0 to 50 (mV)” to “0 to 100 (litter/min)

This is the actual scale (physical quantity) programming for the printing range being programmed in the [Range/Printing range]. 2 Printing range will be copied

When programming the [Range/Printing range], the printing range is copied to the programming value of the scale.

If the scale copied from the printing range is used, the decimal point positions programmed by the printing range are neglected and the measured values with decimal point positions fixed by range number are displayed. The following table shows the cautions to be observed according to the input types.

0 20 40 60 80 100

Voltage ranges Temperature ranges

Scale programming is not required when the scale is equal to the printing range. They are copied to scale programming, but be careful with the decimal point positions for the scale. See item 3.

Since the printing range is copied to the scale, no programming scale is necessary, but the decimal point position of the measured value is fixed by the range number. For changing the decimal places, refer to 3.

3 Example of Decimal Point Programming

Voltage range Temperature ranges

(1) When the printing range is equal to the scale; (1) To move the decimal point to higher digit Range/Printing

An example of measured values Notes Range/Printing

An example of measured values Notes

0.0 to 50.0 5.00 1 23/0 to 1200 0.0 to 1200..0 850.0 03/0.0 to 50.0 0 to 50 5 23/0.0 to 1200.0 0.0 to 1200.0 850 1 03/0 to 50 0.0 to 50.0 5.0 2 Note 1: Decimal point position of the measured value is

fixed by range number when the printing range is equal to the scale. (2) To move the decimal point to lower digit

by range number when the printing range is equal to the scale. Note 2: Decimal point positions of the programmed scale

An example of measured values Notes

become effective because different decimal place are programmed on printing range and scale. See item (2). (2) When the scale is programmed with a different value from

0 to 200 120.0 25/0 to 200 0 to 200.0 120 1

the printing range. The decimal point position of the programmed scale becomes effective.

Decimal point position rule.

If the decimal point positions of the lower-limit and higher-limit values are different, the lower decimal point position is adopted. 0.00 to 100.0 0.0 to 100.0

For programming the same scale to the other channels;

The [Copy] function is convenient. See Section 11.14.

Show 15 more labels / short text on this page
-69 0 69 · Measuring range · (mV) · Printing range · (mV) · Calibration · Programmed scale · Scale programming · Scale programming · range · range · Range/Printing · Scale programming · range · -57-
Page 62 of the Omega RD200 user manual — ・ CLOCK
Page 62 · click to enlarge
Page 62

11. OTHER PROGRAMMING 11.2 Scale

4 Programming Mode

Scale (max. 11 digits by left-justify) Unused digits are filled with spaces

Min. value (within 5 digits) to Max. value (within 5 digits) Scale value to

print 0% + Scale value to

SHIFT - (-)

To clear (none) the scale programmed

5 Programming Flow Chart

–100 to +100 for CH 3

(1) Select the channel to program to

“none” with or .*

SHIFT 0 CLEAR

simultaneously to clear and execute

SHIFT 4 SCALE

Ref. 1

[temporarily storing] and [storing]. * If a channel is programmed by a numeric value, not by up and down keys, and cleared, the channel before programming change is programmed to scale “none”. Note) The scale of the channel, of which

[To be ready for programming]

SET END

3 UNIT

scale was programmed to “none”, is reflected by the printing range being programmed in the [Range/Printing range]. Tag and alarms are cleared (not programmed). Press or to check the programmed scale of other channels. • Adding/deleting decimal point Move the cursor to the right:

Ref. 2

Other channels check Reference 1

YES

Any key other than SHIFT

NO

YES

  • or → Numeric keys
  • Programming “to” : +
  • Space: Store the [temporarily stored] parameters into the memory. (A programming change mark is printed.) * Fill unused digits with spaces.

・ CLOCK

SPACE COPY(=)

SHIFT - (-) SPACE COPY(=)

NO

SHIFT SET END

Ref. 3

Show 24 more labels / short text on this page
Character display · Channel · print 100% · % / · (2) Press and · + · [Check] · Channel · [Selecting channel] · Channel · [Programming scale] · Various · keys · Scale ※ · / % · Error message appears? · Temporarily stored · Reference 2 · Various keys · Programming other channels? · + · Storing Reference 3 · Stored · -58-
Page 63 of the Omega RD200 user manual — SHIFT 0 CLEAR
Page 63 · click to enlarge
Page 63

11. OTHER PROGRAMMING 11.3 Skip (Channel Deletion)

When “CLEAR” (non-programming condition) is stored in the [Range/Printing range] programming, printing and display of specific channels are skipped. Press and in the operation screen to display the [Range/Printing range] programming screen. The default is no skip function programmed to any channel. 1 Skipping

SHIFT 7 RANGE

Your recorder functions assuming that these channels do not exist. (1) Measured value: Disappear (2) Bargraph: Disappear (3) Data printing: Disable (4)Trace printing: Overshooting to the minimum limit. 2 Programming Skip Flow Chart

Operation of skipped channels

If the skip function is programmed to the specified channel, programmed parameters of [scale], [alarm], [engineering unit] and [tag] of this channel are also cleared. Program the parameters again if the channels are restored to no-skipped status.

Select a channel with or . If a channel is selected with a numeric key and the “CLEAR” is executed, the channel being displayed before changing programming ([1] in the flow chart) is deleted. You can check the range programming of other channels. Press or to change channels. The skipped channels are not displayed. Store the [temporarily stored] skip into the memory. (A programming change mark is printed.)

Skipping CH 3 (3rd pen)

SHIFT 7 RANGE

Ref. 1

SET END

[To be ready for programming]

or (Press several times)

Other channel check Reference 1

SHIFT 0 CLEAR

YES

NO

SHIFT SET END

Ref. 2

Show 18 more labels / short text on this page
To select channels · / % · + · [Check] · Channel · % / · [Selecting channel] · / % · Channel · + · [Programming clear] · Channel · Storing Reference 2 · Temporarily stored · Skipping other channels? · + · Stored · -59-
Page 64 of the Omega RD200 user manual — 3 New Programming Flow Chart of Skipped Channel
Page 64 · click to enlarge
Page 64

11. OTHER PROGRAMMING 11.3 Skip (Channel Deletion)

3 New Programming Flow Chart of Skipped Channel

To the range of 0 to 1200 °C (K) for skipped channel (CH 3)

The programmed parameters of [scale], [alarm], [engineering unit] and [tag] are cleared at the skipped channels. 1) Select a computing unit in the [°C/°F computing selection] if the channels are newly programmed to temperature range. 2) Program the required [scale] if the channels are newly programmed to voltage range. 3) Program [alarm], [engineering unit] and [tag] if necessary. You can check the range programming of other channels. Press or to change channels. The skipped channels are not displayed. This programming can be started at any channel. With the example flow chart, the programming starts from CH 1. As the programmed parameters of CH 1 are not changed, they are same as original parameters. In the example, CH 3 is a channel to be newly programmed. If CH 3 has not been programmed as a skip channel, new range/printing range is displayed. Refer to Section 8.2. Store the [temporarily stored] parameters into the memory. (A programming change mark is printed.) * Fill used digits with spaces.

Omega RD200 — 3 New Programming Flow Chart of Skipped Channel diagram (page 64)
Fig. 192 · 3 New Programming Flow Chart of Skipped Channel

Skipped channel check Reference 1

Starting channel Reference 2

Channel No. for new programming Reference 3

Range/Printing range programming Reference 4

Show 4 more labels / short text on this page
Other programming parameters · / % · Storing Reference 5 · -60-
Page 65 of the Omega RD200 user manual — SHIFT - (-)
Page 65 · click to enlarge
Page 65

11. OTHER PROGRAMMING 11.4 Subtract Printing

This programming is for printing a difference between channels or between a channel and a reference value. Press and simultaneously to display the [Range/Printing range] programming screen to program the subtract printing. 1 Subtract Printing Types

SHIFT 7 RANGE

Subtract printing can be specified to any channel. For example, [CH 1 minus CH 2] can be specified to [CH 2]. In this case, the difference between channels are displayed and printed at CH 2.

Two types of subtract printings are provided. (1) Printing a difference between a reference channel and a subtraction channel (2) Printing a difference between a reference channel and a specified subtraction value (reference value).

Concept of subtract printing

Programming reference channel and subtraction channel

Printing the difference between CH 1 and CH 2 at CH 3

Be sure to program [Range/Printing range] of both the reference channel and the subtraction channel in advance.

CH 1

0 500 ºC

Place a decimal point to the reference value

CH 2

0 500 ºC

Program the reference value with the scale programmed value within 5 digits. For identifying the reference value to a channel No., be sure to place a decimal point to the reference value having no decimal point.

CH 3

100 -250 250ºC Printing range

2 Programming Modes

1. Mode 1

“R” stands for the subtract printing mode 1.

Unused digits are filled with spaces.

Subtraction channel or reference value (within 5 digits)

2. Mode 2 (To move from Mode 1 to Mode 2, press after completing the programming for the Mode 1.)

“D” stands for the subtract printing mode 2.

(11 digits by left justify)

Unused digits are filled with spaces.

Min. value (within 5 digits) to Max. value (within 5 digits) Scale value to

print 0% + Scale value to

SHIFT - (-)

Show 20 more labels / short text on this page
Subtract printing channel · Character display · Programming lamp · Character · display · Subtract · printing · Equal · Reference · channel · Minus · Memory colon · Character display · Programming lamp · Character · display · Subtract printing range · Memory colon · print 100% · -61-
Page 66 of the Omega RD200 user manual — NO
Page 66 · click to enlarge
Page 66

11. OTHER PROGRAMMING 11.4 Subtract Printing

3 Programming Flow Chart

Printing [Channel 1 – Channel 2] at Channel 3 within a printing range of ±250

Other channels check Reference 1

You can check the range programming (including subtract printing) of other channels. Press or to change channels.. If a specified value (reference value) is programmed instead of the subtraction channel in the flow chart, the difference with the reference value will be printed. Add a decimal point to the reference value.

SHIFT 7 RANGE

Ref. 1

[To be ready for programming]

SET END

Programming reference channel and subtraction channel Reference 2

3 UNIT

[Programming channel for subtract printing]

SHIFT /

1 CHART

Various keys Reference 3

  • Programming “to”: +

SHIFT - (-) ・ CLOCK

- (-) 2 TAG

  • Decimal point: in the next digit

[Programming subtraction channel] →

Ref. 2

  • Deletion of decimal point: in

SPACE COPY(=)

the next digit Store the [temporarily stored] parameters into the memory. (A programming change mark is printed.) * Fill unused digits with spaces.

[Programming subtract printing range]

Ref. 3

Subtract printing range ∗

YES

Any key other than SHIFT

NO

YES

NO

SHIFT SET END

Ref. 4

Show 25 more labels / short text on this page
Reference · Subtraction · Subtract printing · channel · channel · channel · + · [Check] · / % · Channel · Channel · [Programming equal] + · [Programming reference channel] · Reference channel · Subtraction channel Minus · [To Mode 2] · Storing Reference 3 · Various · keys · Error message appears? · Temporarily sotred · Programming other channels? · + · Stored · -62-
Page 67 of the Omega RD200 user manual — D
Page 67 · click to enlarge
Page 67

11. OTHER PROGRAMMING 11.5 Alarm

Alarm parameters (alarm types, alarm value, etc.) can be programmed for each alarm point (channel, level). By programming the alarm parameters, alarm activation can be displayed and also the activation and reset can be printed. See Section 10.3. Press and simultaneously to display the [Alarm] programming screen. Alarm output is an option. The alarm outputs for “FAiL (failure) ” and “C. End (chart paper end)” are available in addition to measured values. 1 Alarm Parameters The default is no alarm programmed. 1) Alarm points (Channel, level) Alarm point can be programmed up to level 4 for each channel. “FAiL” and “C. End” can be programmed instead of the channels.

SHIFT 8 ALARM

4. Others Rate-of-change alarm: Program measuring

count (1 to 20) Differential alarm: Program channels to be compared. (1) The above figure shows programming alarm

FAiL…Alarm when the hardware except servo-circuit/mechanism is abnormal In this condition, the status lamp is not lit. C. End…Alarm at just before when the chart paper

Relations between alarm parameters and alarm activation

ends.

Channel Level Alarm type (Relay No.)*

2. Alarm types 10 different alarm types can be selected for each alarm point.

50 0 H

60 0 H

70 0 H

: High alarm : Low alarm Absolute

: Low alarm with standby Rate-of-

valueHigh alarm with standby
changeIncrease limit : Decrease limit

: Differential low alarm Differential : Differential high

: Differential high alarm

point up to level 4 on CH 1. (2) Alarm is in active at level 2 (high alarm) as the

: Differential low alarm with standby

measured value (55) is higher than the alarm value (50). *No alarm output is provided to the standard specifications and the relay No. is fixed to “0”. Refer to Section 12.2 for the programming of relay No. 2 Alarm Types

3. Alarm value Value for alarm activation

The high alarm for the absolute value activates when the measured value equals to or is higher than the alarm value. The low alarm activates when the measured value equals to or is lower than the alarm value. For the rate-of-change alarm and the differential alarm, refer to the followings.

This is the function not to activate an alarm until the alarm becomes in reset condition, even if the alarm is in active condition. This function is only available when the power supply is turned on. When changing parameters, this function is only available by reprogramming it after clearing once. Difference of measured value (absolute value) equals to or is higher than alarm value: Differential high alarm activation Difference of measured value (absolute value) equals to or is lower than alarm value: Differential low alarm activation

Increase limit alarmChange width (PV2 to PV1) per unit

time (∆t) is plus. Decrease limit alarm: Change width (PV2 to PV1) per unit

time (∆t) is minus.

(Differential high alarm) (Differential low alarm)

(PV2 to PV1) Difference of measured

D

value Difference of measured

High alarm Low alarm

Unit time (∆t) = Measuring interval (about 0.1 sec.) x

Measuring count (1 to 20) * Do not apply a “minus” symbol to the alarm value of the decrease limit alarm ().

Show 24 more labels / short text on this page
Absolute value alarm · Alarm points · Alarm specifications · Alarm value · 0 100 (55) · Measured value · alarm with standby · Alarm with standby · Rate-of-change alarm · alarm value U · + · Differential alarm · Change width · △t · Alarm · alarm value* · value · value (absolute value) · Alarm · value (absolute value) · - · Alarm activation · Alarm activation · -63-
Page 68 of the Omega RD200 user manual — 3 Programming Mode
Page 68 · click to enlarge
Page 68

11. OTHER PROGRAMMING 11.5 Alarm

3 Programming Mode

1. Absolute value alarms ( , , , )

Relay No.* Alarm value (Note) Max. 5 digits by left-justify Unused digits are filled with spaces.

(Note) Program the alarm value within the scale range.

2. Rage-of-change alarm ( , )

Relay No.*

Alarm value (Note 1) Max. 5 digits by left-justify Unused digits are filled with spaces.

Measuring count (Note 2) (1 to 20)

(Note 1) Program the alarm value with the change width (without sign) per unit time. (Note 2) Unit time = measuring interval (about 0.1 sec.) x measuring count

Program this measuring count (1 to 20).

3. Differential alarm ( , , , )

Relay No.*

Alarm value (Note 1) Max. 5 digits by left-justify Unused digits are filled with spaces.

Channel to be compared (Note 2)

(Note 1) Program an alarm value with a difference (without sign) between measured values of channels. (Note 2) Program another channel to be compared.

* In the standard specifications, the relay No. is fixed to “0” and the cursor does not appear. Program the relay No. for the alarm output (option) only. See Section 12.2.

Show 17 more labels / short text on this page
Character · display · Channel · Level Alarm · type · Character · display · Cursor Cursor · Channel · Level · Alarm type · Character · display · Channel · Alarm type · Level · -64-
Page 69 of the Omega RD200 user manual — 4 Programming Flow Chart
Page 69 · click to enlarge
Page 69

11. OTHER PROGRAMMING 11.5 Alarm

4 Programming Flow Chart

Follow the [Programming alarm point] procedure and select the alarm point to be deleted with . Clear it by pressing and simultaneously and perform [temporarily stored] and [stored] operation. By pressing , the level advances from 1 to 4 and the channel advances to the next channel and then the level of the next channel advances. Press for reverse operation. 1. Channel can be selected also with or . 2. Channel advances by pressing . Then “FaiL” and “C.End” appear but these functions are only available in the alarm output (option). Relay No. is fixed to “0” and no cursor appears as the alarm output (option) is not added. The relay No. for AH is displayed with 2 digits of “00”. Programming range is 1 to 20. Reference channel to be compared is programmed in 1 to 4. (The differential alarm cannot be selected in a 1-pen type recorder because no channel to be compared is existed.) Store the [temporarily stored] parameters into the memory. (A programming change mark is printed.)

Omega RD200 — 4 Programming Flow Chart diagram (page 69)
Fig. 193 · 4 Programming Flow Chart
1. Without alarm output (option)
Omega RD200 — 4 Programming Flow Chart diagram (page 69)
Fig. 194 · 4 Programming Flow Chart

Other alarm point check Reference 1

Programming alarm point Reference 2

Relay No. Reference 3

Cursor appears only for rate-of-change alarm. Reference 4

When error appears press any key other than SHIFT

Cursor appears only for differential alarm. Reference 5

Show 7 more labels / short text on this page
/ · % · % · / · / · Storing Reference 6 · -65-
Page 70 of the Omega RD200 user manual — YES
Page 70 · click to enlarge
Page 70

11.OTHER PROGRAMMING 11.5 Alarm

2. With alarm output (option)

Follow the [Programming alarm point] procedure and select the alarm point to be deleted with .

To delete alarm point

SHIFT 8 ALARM

Ref. 1

Clear it by pressing and

SHIFT 0 CLEAR

simultaneously and perform [temporarily stored] and [stored] operation. The relay Nos. in alarm output are displayed. This will be blank unless all of the relays are operated. By pressing , the level advances from 1 to 4 and the channel advances to the next channel and then the level of the next channel advances. Press for reverse operation. 1. Channel can be selected also with or . 2. Channel advances by pressing Then “FaiL” and “C.End” appear. If these functions are selected, program relay Nos. only. This is for specifying terminal Nos. for alarm output. The relay No. for AH is displayed with 2 digits such as “00”. See Section 12.2.When programming “1” to “9”, two entering methods of 1 to 9 or 01 to 09 are available. ( : space) Programming range is 1 to 20. Reference channel to be compared is programmed in 1 to 4. (The differential alarm cannot be selected in a 1-pen type recorder because no channel to be compared is existed.) Store the [temporarily stored] programmed values into the memory. (A programming change mark is printed.)

Relay No.

Ref. 2

Output check Reference 1

SET END

[To be ready for programming]

Other alarm point check Reference 2

1 CHART 4 SCALE to

Ref. 3

% / ° F

Programming alarm point Reference 3

Ref. 4

[Programming relay No.]

Relay No.

Relay No. Reference 4

Ref. 5

° F

° F

Ref. 6

Cursor appears only for rate-of-change alarm. Reference 5

Channel to be compared

When an error appears press any key other than SHIFT

Cursor appears only for

differential alarm. Reference 6

Programming other alarm points?

YES

NO

SHIFT SET END Storing Reference 7

Ref. 7

Show 29 more labels / short text on this page
/ · + · [Output check] · [Programming check] · / · [Programming alarm point] · Channel · Level · → · % · [Selecting alarm type] · Alarm type · Numeric keys · / % · / · Numeric keys · [Programming alarm value] · Alarm value · → · Numeric keys · [Only rate-of-change alarm] · Measuring count · → · Numeric keys · [Only differential alarm] · Temporarily stored · + · Stored · -66-
Page 71 of the Omega RD200 user manual — ° F 5 DATA I
Page 71 · click to enlarge
Page 71

11. OTHER PROGRAMMING 11.6 Alarm Deadband

A deadband can be programmed between alarm-activation and alarm-reset. Press and simultaneously in the operation screen for 3 seconds or more to display the [Alarm deadband] programming screen. This programming is common to all alarm points. 1 Alarm Deadband

SHIFT ° C

  • An alarm activates when a measured value exceeds the alarm value. The alarm reset is executed at a value that is lower than the alarm value. This difference is called the deadband and is specified with a scale width (%).
  • The programming range is 0.1 to 9.9% and can be programmed in 0.1% increments. The default is 0.1%.

Low alarm High alarm

OFF OFF

2 Programming Flow Chart

MIN MAX MAX MIN

Alarm value Alarm value

From 0.1% to 0.5%

3 sec. or more

SHIFT

° C

Decimal point is fixed.

SET END

[To be ready for programming]

° F 5 DATA I

Show 15 more labels / short text on this page
Alarm deadband · Dead · Dead · band ON ON · band · (Activation) · (Activation) · (Reset) (Reset) · ( + ) · [Check] · , · [Programming deadband] · Deadband · Stored · -67-
Page 72 of the Omega RD200 user manual — NO
Page 72 · click to enlarge
Page 72

11. OTHER PROGRAMMING 11.7Periodic Data Printing

This programming is for digital printing (data printing) at fixed intervals. The printing overlaps with trace printing. Press and simultaneously to display [Periodic data printing] screen. As the default is no programming of parameters for the periodic data printing, the periodic data printing is not executed. 1 Periodic Data Printing

SHIFT 5 DATA I

Omega RD200 — SHIFT 5 DATA I diagram (page 72)
Fig. 195

(1) Program start time and interval time. (2) The printing format with a chart speed of 99 mm/h or

slower is shown below. 18:30 1: 225°C 2: 316°C 2 Programming Mode

Omega RD200 — SHIFT 5 DATA I diagram (page 72)
Fig. 196

3 Programming Flow Chart

Interval time Start time

It depends on the chart speed and number of printing lines.

Shortest time of interval time (T)

Periodic data printing with a start time of 12:00 and

interval time of 4 hours

180 x number of printing lines* T (Min.) ≧ Chart speed *1 (Note) However, T is 5 minutes or more. *1: When 3-speed is programmed by the remote contacts (option), the slowest speed is applied. *2: Depending on the chart speed and number of channels. Chart speed CH1 CH2 CH3 CH4 99 mm/h or slower 1 line 2 lines 100 mm/h or faster 1 line 2 lines 3 lines 4 lines

SHIFT 5 DATA I

[To be ready for programming]

SET END

0 CLEAR

4 SCALE

Press and simultaneously to clear and carry out [store]. The programmed values in the [Periodic data printing] are cleared and the periodic data printing stops. Reprogram it if necessary. In case electric power failure occurs and the power is turned on in the following

To program periodic data printing to “none”

0 CLEAR 0 CLEAR

SHIFT 0 CLEAR

1 CHART

2 TAG

0 CLEAR 0 CLEAR

When the chart speed is changed;

From power off to on

YES

Any key other than SHIFT

NO

day, reprogramming is required if 24

not an integer. (TInterval time)
Show 14 more labels / short text on this page
Time · (Fixed) (Fixed) · + · [Check] · → → → · [Programming interval time] · Interval time · → → → · [Programming start time] · Start time · Error message appears? · T is · Stored · -68-
Page 73 of the Omega RD200 user manual — ° F
Page 73 · click to enlarge
Page 73

11. OTHER PROGRAMMING 11.8 Engineering Units

Engineering units up to 5 digits can be assigned for digital data printing and scale printing. Press and simultaneously to display [Engineering unit] programming screen.

SHIFT 3 UNIT

(Digital data printing) (Scale printing) (List printing) 1 In case no engineering unit is programmed;

単位印字の例 Examples of engineering unit printing

Omega RD200 — SHIFT 3 UNIT diagram (page 73)
Fig. 197
Omega RD200 — SHIFT 3 UNIT diagram (page 73)
Fig. 198
Omega RD200 — SHIFT 3 UNIT diagram (page 73)
Fig. 199

An engineering unit is decided by the range number programmed in the [Range/Printing range] programming.

Voltage range Temperature range

°C (Nos. other than ones

mV (01 to 05)

V (06 to 10)

K (47, 80)

shown on the right)

2 Programming Mode

(Up to 5 digits.) (Fixed)

Digit No. of engineering unit character *2

(Numeric numbers (1 to 5) appear for the digit not programmed by pressing .)

° F

*1: Engineering unit character The character at the digit where the digit number is pointed by the cursor is displayed. *2: Digit No. of engineering unit character A digit No. 1 to 5 is displayed by pressing . Press and simultaneously to clear the digit No. and press for reprogramming a new digit No.

° F

° F SHIFT 0 CLEAR

If clear is stored at scale programming; If the range No. is changed;

A programmed engineering unit is deleted and the engineering unit determined by the range No. is displayed

If the scale is cleared in a channel; the engineering unit of this channel is deleted, and it becomes the engineering unit determined by range No.

Show 5 more labels / short text on this page
Engineering · unit character · Channel · *1 · -69-
Page 74 of the Omega RD200 user manual — SPACE COPY(=)
Page 74 · click to enlarge
Page 74

11. OTHER PROGRAMMING 11.8 Engineering Units

3 Programming Flow Chart

(1) Select the channel to be “none” with or .*

To program engineering unit to “none”

From PPM to G/MIN in CH 3

(2) Press and simultaneously to clear and perform

SHIFT 0 CLEAR

SHIFT 3 UNIT

Ref. 1

[temporarily stored] and then [stored]. * If the channel is programmed by a numeric value and cleared, the engineering unit of channel before programming change is programmed to be “none”.

SET END

[To be ready for programming]

To check all programmed digits Reference 1

  • Press or to change channels.
  • The digit displaying an engineering unit character (digit No. with decimal point) shifts with or .

3 UNIT

° F

[Programming engineering unit character]

Ref. 2

° C ° F

Programmable characters and key operation Reference 2

  • Numeric value (0 to 9):

0 CLEAR 9 LIST - (-)

  • Alphabetic characters (A to Z):

SHIFT 6 A ~ Z

simultaneously and select a character

YES

by pressing or .

NO

  • Special characters (%, /, ° C, ° F)

% ( + )

SHIFT

SHIFT SET END

/ ( + )

SHIFT

/ ° C

Ref. 4

° C ( + )*

SHIFT

° F ( + )*

° F

SHIFT

*A space corresponding to 2 digits is used. • (space): A space is treated as a character. These engineering units are treated as a printing unit only. They are not treated as computing units. Store the [temporarily stored] parameters into the memory. (A programming change mark is printed.)

SPACE COPY(=)

°C, °F engineering unit Reference 3

Show 23 more labels / short text on this page
% / · + · [Check] · Engineering unit character · Channel · Decimal point · [Programming channel] · % / · Engineering unit character · Channel · , · Various keys · Engineering unit character · to , and · Temporarily stored · Press and · Programming other channels? · % / · + · % · Stored · Storing Reference 4 · -70-
Page 75 of the Omega RD200 user manual — ° F
Page 75 · click to enlarge
Page 75

11. OTHER PROGRAMMING 11.9 Tags

Tags up to 9 digits can be assigned for digital data printing and scale printing. Press and simultaneously to display the [Tag] programming screen. The default is no programming of tags.

2 TAG

SHIFT

(Scale printing) (List printing) 1 Programming Mode

Example of tag printing (RD200)

Omega RD200 — 2 TAG diagram (page 75)
Fig. 200

(Up to 9 digits.) (Fixed)

Digit No. of tag character *2

° F

(Numeric numbers (1 to 9) appear for the digit not programmed by pressing .)

*1: Tag character The character at the digit where the digit number is pointed by the cursor is displayed. *2: Digit No. of tag character A digit No. 1 to 9 is displayed by pressing . Press and simultaneously to clear the digit No. and press for reprogramming a new digit No.

° F

° F SHIFT 0 CLEAR

If clear is stored in the scale mode;

If the scale is cleared in a channel; the tag of this channel is deleted. Reprogram it if necessary.

Show 4 more labels / short text on this page
Tag character · Channel · *1 · -71-
Page 76 of the Omega RD200 user manual — SPACE COPY(=)
Page 76 · click to enlarge
Page 76

11. OTHER PROGRAMMING 11.9 Tags

2 Programming Flow Chart

(1) Select the channel to be “none” with or .* (2) Press and

From TIC to 10 in CH 3

To program tag to “none”

SHIFT 2 TAG

Ref. 1

SHIFT 0 CLEAR

  • simultaneously to clear and perform [temporarily stored] and then [stored]. * If the channel is programmed by a numeric value and cleared, the engineering unit of channel before programming change is programmed to be “none”. Tag characters are displayed only for the programmed channels (Digit Nos. appear.).
  • Press or to change channels.
  • The digit displaying a tag character (digit No. with decimal point) shifts with or .

SET END

[To be ready for programming]

3 UNIT

To check all programmed digits Reference 1

° F

Ref. 2

° C ° F

Programmable characters and key operation Reference 2

YES

  • Numeric value (0 to 9):

0 CLEAR

9 LIST - (-)

NO

  • Alphabetical characters (A to Z):

SHIFT SET END

SHIFT 6 A ~ Z

simultaneously and select a character

Ref. 3

by pressing or .

  • Special characters (%, /, ° C, ° F)

SHIFT

% ( + )

/ ( + )

SHIFT

/ ° C

° C ( + )*

SHIFT

° F ( + )*

° F

SHIFT

*A space corresponding to 2 digits is used. • (space): A space is treated as a character.

SPACE COPY(=)

Store the [temporarily stored] parameters into the memory. (A programming change mark is printed.)

Show 21 more labels / short text on this page
% / · + · [Check] · Channel · [Programming channel] · Channel Tag character · , · Various · keys · [Programming tag character] · % / · Temporarily stored · Programming other channels? · + · to , and · Press and · Stored · % / · % · Storing Reference 3 · -72-
Page 77 of the Omega RD200 user manual — ° F
Page 77 · click to enlarge
Page 77

11. OTHER PROGRAMMING 11.10 Message

Messages are printed from (1) key operation or (2) remote contacts signal*. A message up to 15 digits and 5 different messages (No.1 to No. 5) can be programmed. Press and simultaneously for 3 seconds or more to display the [Message] programming screen. * This function is only available in the remote contacts (option). See Section 13.1.

0 CLEAR 6 A ~ Z

Example of message printing

Omega RD200 — 0 CLEAR 6 A ~ Z diagram (page 77)
Fig. 201

1 Programming Mode

programmed for No.1.

(Up to 15 digits.) (Fixed)

No. (1 to 5)

Digit No. of message character *2

° F

(Numeric numbers (1 to 15) appear for the digit not programmed by pressing .)

*1: Massage character The character at the digit where the digit number is pointed by the cursor is displayed. *2: Digit No. of message character A digit No. 1 to 15 is displayed by pressing . Press and simultaneously to clear the digit No. and press for reprogramming a new digit No.

° F

° F SHIFT 0 CLEAR

Show 4 more labels / short text on this page
Message · character · *1 · -73-
Page 78 of the Omega RD200 user manual — SPACE COPY(=)
Page 78 · click to enlarge
Page 78

11. OTHER PROGRAMMING 11.10 Message

2 Programming Flow Chart

(1) Select the channel to be “none” with or .* (2) Press and

To program message to “none”

From “none” to “MOTOR ON”.

SHIFT 0 CLEAR

( + ) 3 sec. or more

0 CLEAR

6 A ~ Z

  • simultaneously to clear and perform [temporarily stored] and then [stored]. * If the channel is programmed by a numeric value and cleared, the engineering unit of channel before programming change is programmed to be “none”. Message characters are displayed only for the programmed channels. (Digit Nos. appear.).
  • Press or to change channels.
  • The digit displaying a tag character (digit No. with decimal point) shifts with or .

Ref. 1

No.

SET END

[To be ready for programming]

To check all programmed digits Reference 1

1 CHART

[Programming No.]

No. Message character

° F

Ref. 2

° C ° F

Programmable characters and key operation Reference 2

  • Numeric value (0 to 9):

0 CLEAR

9 LIST - (-)

YES

Programming other Nos.?

  • Alphabetical characters (A to Z):

SHIFT 6 A ~ Z

NO

simultaneously and select a character

SHIFT SET END

by pressing or .

  • Special characters (%, /, ° C, ° F)

Ref. 3

% ( + )

SHIFT

/ ( + )

SHIFT

/ ° C

° C ( + )*

SHIFT

° F ( + )*

SHIFT

° F

*A space corresponding to 2 digits is used. • (space): A space is treated as a character. Store the [temporarily stored] parameters into the memory. (A programming change mark is printed.)

SPACE COPY(=)

Show 15 more labels / short text on this page
% / · [Check] · % / · , · Various · keys · Temporarily stored · to , and · + · Press and · % / · Stored · % · Storing Reference 3 · -74-
Page 79 of the Omega RD200 user manual — YES
Page 79 · click to enlarge
Page 79

11. OTHER PROGRAMMING 11.11 Burnout

The channel where the temperature range is programmed becomes effective. Press and simultaneously in the operation screen for 3 seconds or more to display the [Burnout] programming screen. This programming is necessary for each channel. 1 Burnout

SHIFT %

Burnout can be selected for each channel out of 3 types shown below.

  • If a sensor (thermocouple or resistance thermometer) is disconnected, trace printing overshoots maximum or minimum limit.
  • The default parameter of burnout is “disable” ( ) in all channels.
(Non)Burnout disable (UP burn) : Up scale burnout (down burn) : Down scale burnout

2 Programmed Flow Chart

A parallel connection of a thermocouple with other instrument causes a problem. Select () if your recorder uses the thermocouple being connected with other instrument. Even if the burnout is programmed to a channel where the voltage range (No. 01 to 07) is programmed, this programming becomes ineffective (burnout disable). For the burnout operation (display and printing), read Section 10.2. Press or to change channels. The programmed burnout of other channels can be checked. Store the [temporarily stored] parameters into memory. (Programming change mark is printed.)

From burnout “disable” to “up scale burnout”

“” programming on parallel operation

( + ) 3 sec. or more

SHIFT

Ref. 2

SET END

[To be ready for programming]

Voltage range is ineffective

2 TAG

Press several times.

Burnout operation Reference 1

Other channels check Reference 2

YES

NO

SHIFT SET END

Ref. 3

Show 14 more labels / short text on this page
Burnout selection menu · % · [Check] · Channel · [Specifying channel] · / · [Selecting burnout] · Temporarily stored · / % · Programming other channels? · + · Storing Reference 3 · Stored · -75-
Page 80 of the Omega RD200 user manual — SHIFT
Page 80 · click to enlarge
Page 80

11. OTHER PROGRAMMING 11.12 Passcode/Key Lock

The procedure for [Key lock] programming differs depending on whether a passcode is programmed or not. Press and simultaneously in the operation screen for 3 seconds or more to display the [Passcode programmed/not programmed] check screen. If a passcode has been already programmed, it is not allowed to go to the [Key lock] programming screen unless entering the correct passcode. 1 Passcode

SHIFT /

3 Programming Flow Chart

  • [Key lock] programming cannot be allowed unless entering the correct passcode.
  • As the default is no passcode programmed, you can program [Key lock].

The flow differs depending on whether a passocde is programmed or not. See next page for details.

( + )3 sec. or more

SHIFT /

YES

Program a passcode with a 4-digit numeric value. Programming range: 0001 to 9999 If “0000” is programmed, a programming error occurs.

NO

PASS CODE non PASS CODE

0 CLEAR

A Keep your passcode in mind

9 LIST

Keep the passcode programmed in mind or somewhere safe. If you lose the passcode, no [Key lock] operation will be available. 2 Key Lock

NO

YES

SET END

PASS CODE PASS CHANGE LOCK

When [Key lock] is programmed to “LOCKED”, no key operation for the following functions is accepted. (1) Reprogramming of various parameters (2) Operations (printing ON/OFF,

0 CLEAR

A

9 LIST

NO

Program passcode Change passcode?

A

YES

chart paper feeding, digital data printing, printing format selection, and time axis synchronization selection) The following operations are exceptional. (1) Checking various parameters (2) Programming the key lock to be

KEYLOCK UNLOCKED Select [CHANGE]

[UNLOCK]

KEYLOCK LOCKED Select [LOCK]

ineffective (UNLOCKED) (3) Selection of operation screens

The status KEY LOCK (blue) lights when the key lock is effective.

In key lock Reference

(Note) In case an error message appears,

press any key other than and reprogram.

SHIFT

Show 14 more labels / short text on this page
Passcode programming range · Passcode programmed? · to · Programming · Enter passcode · passcode? · (Note) · to · (Ineffective) · selected · / % · (Effective) · [LOCK] selected · -76-
Page 81 of the Omega RD200 user manual — 4 When a passocde has not been programmed
Page 81 · click to enlarge
Page 81

11. OTHER PROGRAMMING 11.12 Passcode/Key Lock

4 When a passocde has not been programmed

Omega RD200 — 4 When a passocde has not been programmed diagram (page 81)
Fig. 202 · 4 When a passocde has not been programmed

programmed. See Section 5 on the next page. (1) The numeric figures for the passcode

will not appear. (2) If a passcode has been programmed,

the [Key lock] selection is not available without entering the correct passcode. The flow chart shown in 5 is for the procedure of [Key lock] selection when a passcode has been programmed. Press or to select the key lock to be effective or ineffective.

Selection of key lock ineffective or effective Reference 3

-77-

Page 82 of the Omega RD200 user manual — 5 When a passocde has been programmed
Page 82 · click to enlarge
Page 82

11. OTHER PROGRAMMING 11.12 Passcode/Key Lock

5 When a passocde has been programmed

Key lock to be effective when the passcode is changed or unchanged

Omega RD200 — 5 When a passocde has been programmed diagram (page 82)
Fig. 203 · 5 When a passocde has been programmed

[Key check] screen. Press or to select the key lock to be effective or ineffective.

Selection of key lock ineffective or effective Reference 2

-78-

Page 83 of the Omega RD200 user manual — YES
Page 83 · click to enlarge
Page 83

11.OTHER PROGRAMMING 11.13 Input Filter

The input filter has a function to stabilize the measuring input. This function can be programmed for each channel. Press and simultaneously for 3 seconds or more to display the [Input filter] programming screen. 1 Input Filter

0 CLEAR 8 ALARM

Input filter0 (none)

A CR filter is mounted in the measuring circuit. In addition, a software filter (called as “input filter”) for the “primary delay computation” is also installed to smooth slight variations of the measuring input. The value for the programming is corresponding to “Time constant: T”. 2 Programming Flow Chart

A

Input filterT (T = 1 to 10)

Press or to change channels. The input filter time constant of other channels can be checked. 0, 1 to 10 If “0” is programmed, the input filter is disabled. Store the [temporarily stored] parameters into memory. (A programming change mark is printed.)

T

To the time constant 5 seconds in CH2

Other channels check Reference 1

% / 0 CLEAR 8 ALARM

3 sec. or more

Ref. 1

Programming range of time constant

SET END

[To be ready for programming]

2 TAG

° F 5 DATA I

Ref. 2

YES

NO

SHIFT

YES

NO

SET END SHIFT

Show 16 more labels / short text on this page
Input filter · ( + ) · [Check] · Reference 2 · Channel · Time constant · [Programming channel] · Storing Reference 3 · [Programming time constant] · Error message appears? · Keys other than · Temporarily stored · Program other channels? · + · Stored · -79-
Page 84 of the Omega RD200 user manual — 2 Programming Flow Chart
Page 84 · click to enlarge
Page 84

11. OTHER PROGRAMMING 11.14 Copying to Other channels

For the parameters, such as [range/printing range], [scale], [engineering unit] and [tag], which are needed to be programmed in each channel, the parameters of the specific channel as a reference channel can be copied to a desired channel. 1 Programming Mode …Example of [range/printing range]

Omega RD200 — 11. OTHER PROGRAMMING 11.14 Copying to Other channels diagram (page 84)
Fig. 204 · 11. OTHER PROGRAMMING 11.14 Copying to Other channel…

:For sequential channel ・ :For non-sequential channel

* RRange/Printing range S: Scale U: Engineering unit T: Tag

2 Programming Flow Chart

Copying of parameters for [range/printing range] in CH 1 to CH 2 to 4.

Press or to check if any skip channel is programmed.

Omega RD200 — 2 Programming Flow Chart diagram (page 84)
Fig. 205 · 2 Programming Flow Chart

Store the [temporarily stored] parameters into memory. (A programming change mark is printed.) * Fill unused digits with spaces.

Character display · Programming lamp* · -80-

Page 85 of the Omega RD200 user manual — AND/
Page 85 · click to enlarge
Page 85

12. ALARM OUTPUT 12.1 Alarm Output Programming Items

This explanation is only for the alarm output (option). The alarm output programming is necessary after programming the “alarm”. 1 Programming Items for Alarm Output

1. Relay No. Setting of the relay No. where the alarm information (activation/reset) of each alarm point is to be output. Set them to each alarm point.
2. Output wiring Setting to AND or OR for each relay No.. When one relay is used by multiple alarm points, you can select AND or OR for the output wiring.
(1) Relay coil phaseSetting whether N and O terminals are shorted (Energized) or opened

(Non-energized) when an alarm activates.

(2) Relay output latch: Setting whether the alarm status is to be continued until is pressed (Hold) even if the alarm becomes reset condition, or it is reset (Not hold) at the

same time as resetting of the alarm. 3) Output mode

(3) Alarm display* latch: Setting whether the alarm display is kept displayed until is pressed (Hold) even if the alarm becomes reset condition or turned off (Not hold) at the

same time as resetting of the alarm. * Alarm display: ALARM status lamp

2 Details of Programming Items

1. Relay No. and default values Alarm * Alarm point Alarm type Relay No. Level 1 0 Level 2 0 Level 3 0 CH 1
2. AND/OR, output mode and default values Output Mode Relay

AND/

No.

latch RLy1 or RLy2 or RLy3 or RLy4 or RLy5 or RLy6 or RLy7 or RLy8 or RLy9 or RLy10 or RLy11 or RLy12 or

Level 4 0 Level 1 0 Level 2 0 Level 3 0

CH 2 2 to 4-pen types only

Level 4 0 Level 1 0 Level 2 0 Level 3 0

CH 3 3 and 4-pen types only

Level 4 0 Level 1 0 Level 2 0 Level 3 0

CH 4 4-pen type only

* RD200: Relay No. is 1 to 6. RD2800: it is variable depending on the number of output points (Rel;ay No. is 1 to 12 for 12-point outputs.).

Level 4 0 FAiL 0 C. End 0 *This is the “Alarm” programming. The relay No. is fixed at 0 in the standard specifications (AH is fixed at 00). 1 to n (Note) should be set only when alarm output (option) is added. (0 means no output.) (Note) n: 6 or 12 depending on the number of output

points. RD2006 only
Show 15 more labels / short text on this page
Relay output · Alarm display · OR Relay coil · phase · latch · Energized · Not hold · Not hold · Common · Common · Common · to all relays · to all relays · to all displays · -81-
Page 86 of the Omega RD200 user manual — 1 CHART 4 SCALE
Page 86 · click to enlarge
Page 86
12. ALARM OUTPUT 12.2 Programming of Relay No.

This explanation is only for the alarm output (option). Program the relay No. for each alarm point by referring the “Programming relay No.” screen in “Section 11.5 Alarm, 4 Programming flow chart, 2) With alarm output (option)”. 1 Alarm Output Terminals and the Status

N. O COM N. C

When an alarm is in active at an alarm point, the relay No. (alarm output terminal No.) specified for the point is activated.*

N. O COM

*The terminal activation differs depending on the programming of the relay coil to “Energize” or “Non energize”. ⇒ See Section 12.4.

2 Programming Relay No.

For programming [Relay No.], refer to the programming flow chart in “Section 11.5 Alarm, 4 Programming flow chart, 2) With alarm output (option)”. The default is “0” and no alarm output is available. Select an alarm type and press . The cursor moves to the relay No. to specify relay Nos. for each alarm point.

Number of relay No. digits

Each relay No. is composed of two digits. For relay Nos. 0 to 9, set 9 or . Select an alarm type and press to move the cursor to the relay No. screen. No output is available when the relay No. is programmed to “0”. The relay No. for AH is displayed as “00”. Program a relay No. from “0” to “n”. The example is for programming to “3”. Program “0” if no output is necessary. The programming after this step or before the alarm type selection, refer to the flow chart in Section 11.5. 4 2).

1 CHART 4 SCALE

% / ° F

° F

[Programming relay No.]

Relay No.

Number of alarm point and number of output point (n) FAiL, C. End

The total numbers of alarm points are number of channels x number of levels (4) + 2 (FAiL and C. End). Number of output points is 6 or 12 points. (RD200: 6 points only)

Place the cursor to the column of channel in “Selecting alarm type” and press until “FaiL” or “C. End” is displayed. In this condition, select a relay No. to output when the corresponding phenomenon occurs.

Show 16 more labels / short text on this page
Alarm outputs terminals · ~ · [Programming alarm point] · Channel · Level · → · [Selecting alarm type] · Alarm type · Numeric · keys · Numeric · keys · [Programming alarm value] · Alarm value · / · -82-
Page 87 of the Omega RD200 user manual — 7. PROGRAMMING 7.1 Keys and Characters
Page 87 · click to enlarge
Page 87
Omega RD200 — 7. PROGRAMMING 7.1 Keys and Characters diagram (page 87)
Fig. 206 · 7. PROGRAMMING 7.1 Keys and Characters

This explanation is only for the alarm output (option). Press and simultaneously for 3 seconds or more to display the “Output Wiring (AND or OR)” programming screen. Program it for each relay No. The default is “OR” for all relay Nos. 1 AND/OR

Multiple alarm points can be allocated to one relay No.

Omega RD200 — 7. PROGRAMMING 7.1 Keys and Characters diagram (page 87)
Fig. 207 · 7. PROGRAMMING 7.1 Keys and Characters
  • AND output: The relay turns on when all alarm points allocated are in active.
  • OR output: The relay turns on when any of the alarm points allocated is in active. 5 2 Selecting Flow Chart

Press or to change relay No. AND/OR status of other channels can be checked. Store the [temporarily stored] parameters into memory. (Programming change mark is printed.)

From OR to AND in relay No. 3

Omega RD200 — 7. PROGRAMMING 7.1 Keys and Characters diagram (page 87)
Fig. 208 · 7. PROGRAMMING 7.1 Keys and Characters

-83-

Page 88 of the Omega RD200 user manual — 12. ALARM OUTPUT 12.4 Programming Output Mode
Page 88 · click to enlarge
Page 88

12. ALARM OUTPUT 12.4 Programming Output Mode

Omega RD200 — 12. ALARM OUTPUT 12.4 Programming Output Mode diagram (page 88)
Fig. 209 · 12. ALARM OUTPUT 12.4 Programming Output Mode

This explanation is only for the alarm output (option). Press and simultaneously for 3 seconds or

more to display the “Output mode” programming screen. Two output modes, 1) relay coil (energize/not energize),

2. latched alarm display/relay (hold/not hold), are available. The programming is common to all relay Nos.. 1 Relay Coil Energize/Not energize

The default is “Energize”. The terminal configuration differs depending on the type of relay.

1. MOS relay and mechanical relay “a” contact alarm output

-84-

Page 89 of the Omega RD200 user manual — 2 Latched Alarm Display/Relay (Hold/Not hold)
Page 89 · click to enlarge
Page 89

12. ALARM OUTPUT 12.4 Programming Output Mode

2 Latched Alarm Display/Relay (Hold/Not hold)

The default is “Not hold”. The alarm display refers to the blinking of the measured value and the lighting of ALARM status lamp.

1. Not hold ( ) 2) Hold ( )

The output is not affected by .

The output is held until is pressed. The output differs depending

on the timing of pressing .

Blinking of measured value

The measured value blinks when an alarm activates and lights steadily when it is reset. However, the blinking is continued when the alarm display/relay is hold. The blinking is turned to steadily light by pressing regardless of the programming of “Hold” or “Not hold”.

Show 34 more labels / short text on this page
Alarm activation · Alarm activation · Alarm activation · Alarm reset · Alarm reset · Alarm reset · Measured value · Measured value · Measured value · blinking · blinking · blinking · Measured value · Measured value · Measured value · lighting · lighting · lighting · Alarm · Alarm · Alarm lighting · lit · lit · Alarm not lighting · Alarm not lighting · Alarm not lighting · Relay output on · Relay output on · Relay output on · Relay output off · Relay output off · Relay output off · Reference · -85-
Page 90 of the Omega RD200 user manual — 4 Programming Mode
Page 90 · click to enlarge
Page 90

12. ALARM OUTPUT 12.4 Programming Output Mode

4 Programming Mode

Omega RD200 — 4 Programming Mode diagram (page 90)
Fig. 210 · 4 Programming Mode

Programming the relay coil phase to Energize, relay output/display to Hold and alarm display to Hold:

The parameters are displayed in order of (1) relay coil phase, (2) relay output and (3) alarm display.

Store the parameters into memory. (Programming change mark is printed.)

-86-

Page 91 of the Omega RD200 user manual — CS3 OFF ON
Page 91 · click to enlarge
Page 91

13. REMOTE CONTACTS 13.1 Remote Contacts Function

This explanation is only for the remote contacts (option). 1 Remote Contacts

Omega RD200 — This explanation is only for the remote contacts (option). 1 Remote Contacts diagram (page 91)
Fig. 211 · This explanation is only for the remote contacts (op…

(1) The following functions are available with the contact

signals at remote contacts terminals (EX1 to 4). However, the functions are limited due to four terminals (EX1 to 4) provided. Moreover, some functions are automatically allocated to a certain terminal Nos. (2) Programming to allocate the functions to terminal Nos. is

necessary.

(1) Printing on/off and selection of three chart

speeds 2 terminals (EX1, EX2) Programming 3 speeds, See Section 13.3. (Note 1)

(2) Execution of message printing (No.1 to 5) 4 terminals (EX1 to EX4) Programming message, See Section 11.10. (Note 2)

(3) Execution of message printing (No.1 and 2) 2 terminals (EX1, EX2) Programming message, See Section 11.10. (Note 3)

(4) Execution of digital data printing Any 1 terminal

(5) Execution of list printing (list 1, 2, 3) Any 3 terminals

(6) Execution of operation record (No.A to D) Any 4 terminals Programming operation record, See Section 13.4.

(7) Reset of totallization Any 1 terminal Available in totalization (option)

(Note 1) Free terminals left are EX3 and EX4. (Note 2) No free terminals are left. (Note 3) Free terminals left are EX3 and EX4. 2 Functions and Terminal Contact Signals

ONShort OFF: Open

Functions Contact signal at terminal

  • Program 3 chart speeds first. See Section 13.3. With COM

Printing on/off and selection of 3 hart speeds EX1 EX2 CS1 OFF OFF

(1) Printing on/off and selection of

CS2 ON OFF

PrintingOn

CS3 OFF ON

PrintingOff ON ON

(Note) Program “Printing on/off” to on with a key operation in advance.

  • Program messages first. See Section 11.10.

e EX1 EX2 EX3 EX4* No.1 OFF OFF OFF For trigger No.2 ON OFF OFF No.3 OFF ON OFF No.4 ON ON OFF No.5 OFF OFF ON

(2) Execution of message printing

(No. 1 to 5)

1 sec. or more

* When the trigger signals is sent (for 1 sec. or more) after selecting the message No., the printing of programmed message starts. (Note) Program “Printing on/off” to on with a key operation in advance. Execution of message printing with key operation is also available. See Section 9.4.

Show 4 more labels / short text on this page
Functions Terminals Remarks · 3 chart speeds · With COM Messag · -87-
Page 92 of the Omega RD200 user manual — 13. REMOTE CONTACTS 13.1 Remote Contacts Functions
Page 92 · click to enlarge
Page 92

13. REMOTE CONTACTS 13.1 Remote Contacts Functions

ONShort OFF: Open

Functions Contact signal at terminal

  • Program messages first. See Section 11.10.

COM Message EX1 EX2*

(3) Execution of message printing (No.1 and 2)

No.1 OFF

No.2 ON

1 sec. or more

* When the trigger signals is sent (for 1 sec. or more) after selecting the message No., the printing of programmed message starts. (Note) Program “Printing on/off” to on with a key operation in advance. Execution of message printing with key operation is also available. See section 9.4.

Turn on (for 1 second or more) the terminal No. which the digital data printing is allocated.

(Note 1) Program “Printing on/off” to on with a key operation in advance.

(4) Execution of digital data

Execution of message printing with key operation is also available. See Section 9.2. (Note 2) During execution, retry of the

execution can be accepted just once.

1 sec. or more

Turn on (for 1 sec. or more) the terminal No. which the printing of list 1, 2 or 3 is allocated. (Note) Program “Printing on/off” to on with a key

(5) Execution of list printing (List 1, 2, 3)

operation in advance. Execution of message printing with key operation is also available. See Section 9.3.

1 sec. or more

Programming of the operation record position is necessary. See Section 13.4. Turn on (for 1 sec. or more) the terminal No. which the operation record (No. A to D) is allocated. During ON time, the recording position shifts to 5 mm to the right from the programmed operation record position. (Note) Program “Printing on/off” to

(6) Execution of operation record

(No. A to D)

on with a key operation in advance.

Operation record position (10 to 90%)

Available in “Totalization” (option). The totalized value can be reset at an interval time programmed. It can be reset with a contact signal, too.

(7) Reset of totallization

1 sec. or more

Omega RD200 — 13. REMOTE CONTACTS 13.1 Remote Contacts Functions diagram (page 92)
Fig. 212 · 13. REMOTE CONTACTS 13.1 Remote Contacts Functions

For the contact signal applied to the remote contacts terminals, use a switch or a relay driven at 30V AC or less or 60V DC or less or a manual contact for a very light load.

Show 6 more labels / short text on this page
For trigger · printing · ON time · ON time · 5 mm · -88-
Page 93 of the Omega RD200 user manual — + SHIFT SET END
Page 93 · click to enlarge
Page 93
Omega RD200 — 7. PROGRAMMING 7.1 Keys and Characters diagram (page 93)
Fig. 213 · 7. PROGRAMMING 7.1 Keys and Characters

This explanation is only for the remote contacts (option). Press and simultaneously for 3 seconds or more to display the “Terminal Allocation for Operation” programming screen. This allows allocation of the desired functions to terminal Nos. 1 to 4 (EX 1 to 4). 1 Programming Mode

2 Programming Flow Chart

Terminal No. Function

Press or to change terminal Nos. Functions allocated can be checked. The default allocation is as follows:

Checking other terminal Nos.

To allocate a chart speed of 3 and stop to terminal Nos. 1/2 and operation record A to terminal No. 3:

Terminal No. 1Terminal No. 2:

Printing on/off and 3 chart speed selection Terminal No. 3: Digital data printing Terminal No. 4: List printing (List 1)

Ref. 1 3 sec. or more ( + )

0 CLEAR 5 DATA I

With , and , the terminal Nos. are allocated automatically. Therefore, when one of these functions is selected, terminal Nos. 2 to 4 are not displayed. See Section 13.1. 1 .

Terminal Nos. 2 to 4 are not displayed.

SET END

[To be ready for programming]

Ref. 2

Press or to change functions.

[Selecting operation record printing A]

Ref. 3 % /

Displays Details ① None ②

Printing on/off and 3 chart speed selection ③

Message printing (No. 1 and 2) ④

Message printing (No. 1 to 5) ⑤ Operation printing A ⑥ Operation printing B ⑦ Operation printing C ⑧ Operation printing D ⑨ Digital data printing ⑩ List printing (List 1) ⑪ List printing (List 2) ⑫ List printing (List 3) ⑬ Totalization reset

+ SHIFT SET END

Show 13 more labels / short text on this page
Reference 1 · % / · [Check] · Reference 2 · [Selecting operation] · Reference 3 · Types of functions · % · / · [Selecting operation] · Temporarily stored · Stored · -89-
Page 94 of the Omega RD200 user manual — YES
Page 94 · click to enlarge
Page 94

13. REMOTE CONTACTS 13.3 Programming 3 Chart Speeds

The “3 chart speeds” programming screen appears only in the remote contacts (option). Press and simultaneously to display the “3 chart speeds” programming screen. Program 3 chart speeds (CS1 to 3) and select the desired speed with a contact signal. See section 13.1 2. 3 Programming Mode

SHIFT 1 CHART

4 Programming Flow Chart

Chart speed (4 digits programming) Cursor Unit for the

No.

From 20 mm/h to 40 mm/h for CS2

Default: 20 (RD200), 25 (RD2800) Programming range: 0001 to 0600 (mm/h) or 0001 to 0200 (mm/min.) 0000mm/h can be set for RD2800. When 0000mm/h is set, chart is fed at the speed of 12.5mm/h.

SHIFT 1 CHART

Ref. 1

If the chart speed is programmed at 151 mm/h or faster;

Speed No.

SET END

[To be ready for programming]

Ref. 2

All printings except time line, digital data printing and programming change mark will not be executed. See Section 10.1.

% / ° F

[Selecting CS No.]

Influence to periodic data printing

If the chart speed is changed, the parameters programmed in the [Periodic data printing] (Section 11.7) is cleared.

Speed No.

0 CLEAR 4 SCALE 0 CLEAR 0 CLEAR

Other speed No. check Reference 1

YES

Press or to change CS No. Chart speeds can be checked.

Any key other than SHIFT

NO

Changing the speed unit (h/min.)

Place the cursor on the character display (left digit) and press or to switch [M] and [H] alternately.

Programming other Nos.?

YES

NO

SHIFT SET END

Store the [temporarily stored] parameters into memory. (Programming change mark is printed.)

Ref. 3

Show 19 more labels / short text on this page
Speed · speed · Defaults parameters and · programming range · + · [Check] · Speed unit · → · [Selecting speed] · / % · Error display appears? · Temporarily stored · Reference 2 · / · % · + · Storing Reference 3 · Stored · -90-
Page 95 of the Omega RD200 user manual — YES
Page 95 · click to enlarge
Page 95

13. REMOTE CONTACTS 13.4 Programming Operation Record Position

This explanation is only for the remote contacts (option). Press and simultaneously for 3 seconds or more to display the “Operation printing position” programming screen. Printing Nos. are A, B, C and D. The default values are A = 20, B = 40, C = 60 and D = 80. 1 Operation Record

Operation record lines and Nos.

0 CLEAR 2 TAG

B

Record No.

Record line when the contact is open

A

Record line when the contact is shorted

The programmed recording positions are recorded by the plotter pen. When the contact signal is shorted, the recording position shifts to 5 mm to the right from the programmed operation record position. The record Nos. (A to D) are also recorded at a fixed interval. 2 Programming Mode

3 Programming Flow Chart

Record No. Record position

For operation record, the “Terminal Allocation for Operation”, allocations of record Nos. (A to D) to terminal Nos. (EX 1 to EX 4), is necessary. See Section 13.2. Press or to change record Nos. record positions can be checked. Record position is programmable from 10% to 90% of the printing range. Error is displayed when the programmed value is between 0% and 9% or between 91% and 99%. Store the [temporarily stored] parameters into memory. (Programming change mark is printed.)

Setting of the record position of record No. B to 40%

The terminal allocation is necessary.

0 CLEAR 2 TAG

( + ) 3 sec. or more

Ref. 1

Printing No.

SET END

Other speed No. check Reference 1

[To be ready for programming]

% / ° F

[Selecting printing No.]

Printing No.

4 SCALE 0 CLEAR

Ref. 2

If any error occurs, press any key other than .

SHIFT

Ref. 3

YES

NO

SHIFT SET END

Ref. 4

Show 19 more labels / short text on this page
90(% Programming range · 5mm · 5mm · [Check] · Printing position · / % · → · Reference 2 · Record position range · [Programming printing position] · Error display · Printing position · Reference 3 · Temporarily stored · Storing Reference 4 · Programming other channels? · + · Stored · -91-
Page 96 of the Omega RD200 user manual — SET
Page 96 · click to enlarge
Page 96

14. PRINTING FORMAT 14.1 Programming Automatic Range-Shift Printing

This programming display only appears for the printing format (option). This programming is to change the trace printing range automatically. 1 Automatic Range-shift

<Example> Automatic range-shift programming

-200 1370 Measuring value

(R1)

(R4)

The automatic range-shift function changes the trace printing range up to 5 stages according to the measured values. (1) This programming can be applied to each

-100 400 600 800 1200

for 1st range -100 400

channel. (2) The total printing range can be programmed optionally irrespective of the programmed range in the [Range/Printing range]. (3) For each range, refer to the minimum

printing range in section 22.1 (Input Specifications). The printing may be dispersed if each range is programmed to be less than the minimum printing range. (4) The range shifting is executed when the

Mark is the programming point and No. (1) to (5) shows the order of programming. By pressing when the programming ends, the screen moves to the next programming screen. As No. 5 is not used in this example, only press .

measured value exceeds approx. 0.5 mm from the minimum (zero) or the maximum (span) range. 2 Programming Mode

1) Screen 1 (Printing format check)

3) Screen 3 (Zero programming at No.1 range)

Range No. Zero

Dots appear at the selected format. By pressing , the dots disappear and the cursor appears instead. 2) Screen 2 (Programming channel)

(1) The default programmed value is the minimum

Programming value (Max. 5 digits by left-justify)

value in the [Range/Printing range]. This programming can be changed as required. (2) When all programming are completed, return to

SET

END

the above display and execute [Storing]. 4) Screen 4 (Span programming at No. 1 to 5 ranges)

Range No. Span

By selecting automatic range-shift (), the cursor appears at the channel programming position.

(1) When the programming a span, the range

Programming value (Max. 5 digits by left-justify)

number advances. Program the span up to the required range number. (2) After pressing up to range No.5, press again to return to Display 3.

Show 25 more labels / short text on this page
2nd range (R2) · 3rd range (R3) · 1st range · 4th range · Total printing · range · (1) (2) · Printing range · Printing range · (3) · for 2nd range · (400) 600 · Printing range · (4) · for 3rd range · (600) 800 · (5) · Printing range · for 4th range · (800) 1200 · Standard Automatic range-shift · Zone Compressed/expanded · Automatic range selection · Channel Cursor · -92-
Page 97 of the Omega RD200 user manual — SHIFT SET END
Page 97 · click to enlarge
Page 97

14. PRINTING FORMAT 14.1 Programming Automatic Range-Shift Printing

3 Programming Flow Chart Programming the details on the left page for CH 1

Press any key other than and program again. When it is normal, the next screen appears. Program the span and press to advance the range Nos. The range No. advances up to No. 5. For the range No. which no span programming is required, press key to skip it. For storing parameters into memory, return the screen to Step (5). Press to display parameters. The parameters are displayed sequentially each time is pressed. Press or to change channels. Store the parameters into memory. (Programming change mark is printed.) For canceling the stored parameters, press and simultaneously at Steps (4) to (7), and then press to store the cancellation. Point the cursor to () at Step (2) and press . To return to the automatic range-shift function, execute the procedure of Steps (1) to (3) and store it. Scale is printed in the order of channels. Range Nos. (R1 to R5) are also printed. * Fill unused digits with spaces.

Remedial measure to error display

SHIFT

SHIFT DISP

Ref. 1

Programming for the range No. 2 to No. 5

SET END

[To be ready for programming]

[Selecting automatic range shift]

Return to the screen of Step (5)

ENTRY 1 CHART

Checking parameters Reference 1

[Programming zero for range No. 1]

Range No.

[Programming span for range No. 1]

Cancellation of programming Reference 3

Range No. changes to 2. Program the span for the range No. 2. Program No. 2 to 4 spans as the procedure shown above. Press key for every programming.

SHIFT 0 CLEAR

[Returning to the step (5)]

Returning to standard printing Reference 4

SHIFT SET END

Ref. 1

Scale printing at fixed- time printing Reference 5

This programming is for each channel. For programming other channels, repeat the procedure from the [Operation screen].

Show 31 more labels / short text on this page
+ · [Checking printing format] · (1) · Standard · (2) · Automatic range shift · → · (3) · Channel · → · [Programming channel] · (4) · Channel · → · Numeric · keys · (5) · % / · ∗ Zero · → · Numeric · keys · Storing Reference 2 · (6) · Span ∗ · (7) · (8) · + · Stored · Other channel programming · -93-
Page 98 of the Omega RD200 user manual — END
Page 98 · click to enlarge
Page 98

14. PRINTING FORMAT 14.2 Programming Compressed/Expanded Printing

This programming display only appears for the printing format (option). This programming is to print the specified range in the [compressed/expanded] printing mode. 1 Compressed/Expanded printing

<Example> Compressed and expanded programming

-100 600 800 1200

A specified range in the trace printing range can be printed in the compressed or expanded printing mode. (1) This programming can be executed for

20% (No. 1 break point) 0% 100%

80% (No. 2 break point)

each channel. (2) The printing range can be programmed

(2) (3) (4) (5)

optionally irrespective of the programmed range in the [Range/Printing range]. (3) Up to 2 break points can be programmed. 2 Programming Mode

-100 600 800 1200

Mark is the programming point and No. (1) to (5) shows the order of programming. By pressing after the programming ends, the next programming screen appears. (Program No.1 break point and No.2 break point by 2 digits. 5% →05)

1. Screen 1 (Printing format check)

3) Screen 3 (Programming break point printing position)

Dots appear at the selected format. By pressing , the dots disappear and the cursor appears instead. 2) Screen 2 (Programming channel)

(1) Program each break point printing position (%) in

Standard Automatic range shift

Programming No. 1 break point

Programming No. 2

SET

the range of 10 to 90. (2) When all programming are completed, return to

END

the above display and execute [Storing]. (Note) No.1 break point should be lower than No.2

break point. When no programming for the No.2 break point is required, leave it as spaces. 4) Screen 4 (Scale programming of each printing position)

Programming mode of compressed/expanded

By selecting compressed/expanded (), the cursor appears at the channel programming position.

(1) When programming a scale, the printing position

Channel Cursor No. 1 break point

No. 2 break point Programming value (Max. 5 digits by left-justify)

advances to the next position. Program scales up to the 100% position. (2) After completing the programming for 100%,

press to return to Screen 3.

Show 12 more labels / short text on this page
Measuring value · -200 1370 · Com-pre ssed Expanded · Com-pre ssed · Printing range · (1) (1) · Printing position · Scale · Zone Compressed/expanded · break point (Note) · Printing position (%) · -94-
Page 99 of the Omega RD200 user manual — SHIFT SET END
Page 99 · click to enlarge
Page 99
Omega RD200 — 7. PROGRAMMING 7.1 Keys and Characters diagram (page 99)
Fig. 214 · 7. PROGRAMMING 7.1 Keys and Characters

3 Programming Flow Chart Programming the details on the left page for CH 2

Press any key other than and program again. When it is normal, the next screen appears. Leave it as spaces and press . For storing parameters into memory, return the screen to Step (5). Press to display parameters. The parameters are displayed sequentially each time is pressed. Press or to change channels. Store the parameters into memory. (Programming change mark is printed.) For canceling the stored parameters, press and simultaneously at Steps (4) to (7), and then press to store the cancellation. Point the cursor to () at Step (2) and press . To return to the compressed/expanded function, execute the procedure of Steps (1) to (3) and store it. Scales at each printing position are printed like as –100, 600, 800 or1200. • “+” mark is printed at the printing position for break points. * Fill unused digits with spaces.

Remedial measure to error display

SHIFT

SHIFT DISP

No programming of No. 2 break point necessary

SET END

[To be ready for programming]

Return to the screen of Step (5)

Checking parameters Reference 1

No.1 break point

No.2 break point

2 TAG ENTRY

Channel 2 TAG 0 CLEAR 8 ALARM

[Programming printing position for each break point]

0 CLEAR ENTRY →

0 CLEAR

SPACE COPY(=) × 4

[Programming scale of printing position0%]

Cancellation of programming Reference 3

SHIFT 0 CLEAR

Scale * Printing position

Printing position (%) changes. Program the corresponding scale and press for every programming. Continue this programming up to 100% scale.

Returning to standard printing Reference 4

[Returning to the step (5)]

SHIFT SET END

Scale printing at fixed time printing Reference 5

Ref. 2

This programming is necessary for each channel. For programming other channels, repeat the procedure from the “Operation screen”.

Show 25 more labels / short text on this page
+ · Ref1 · [Checking printing format] · (1) · Standard · (2) · Compressed · /Expanded · → · [Selecting compressed/expanded] · (3) · Channel · → · [Programming channel] · (4) · % / · Storing Reference 2 · (5) · (6) · (7) · (8) · + · Stored · Other channel programming · -95-
Page 100 of the Omega RD200 user manual — 4) Screen 4 (Programming channel for zone Nos. 2 to 4)
Page 100 · click to enlarge
Page 100

14. PRINTING FORMAT 14.3 Programming Zone Printing

This programming display only appears for the printing format (option). This programming is to print the printing area by dividing it into 2 to 4 zones. ( RD200: 2 zones only) 1 Zone Printing

Omega RD200 — 14. PRINTING FORMAT 14.3 Programming Zone Printing diagram (page 100)
Fig. 215 · 14. PRINTING FORMAT 14.3 Programming Zone Printing

The printing area is divided into two zones (RD200) or four zones (RD2800) and the printing zone can be selected. This function is effective when printings overlap each other.

(1) The printing zone can be selected for each channel. (2) The printing range for each zone is the programmed

printing range in the [Range/Printing range]. (3) The followings are printing position (mm) at the

number of zone.

  • RD200

Zone No. 1 0 to 45

Zone No. 2 55 to 100

  • RD2800
Omega RD200 — 14. PRINTING FORMAT 14.3 Programming Zone Printing diagram (page 100)
Fig. 216 · 14. PRINTING FORMAT 14.3 Programming Zone Printing

3 0 to 54 63 to 117 126 to 180

4 0 to 36 45 to 81 99 to 135 144 to 180

2 Programming Mode

1) Screen 1 (Printing format check)

3. Screen 3 (Programming channel for zone No.1)

Zone No.1

Dots appear at the selected format. By pressing , the dots disappear and the cursor appears instead. 2) Screen 2 (Programming number of zones) … Displays in RD2800 only

4) Screen 4 (Programming channel for zone Nos. 2 to 4)

Standard Automatic range shift

Cursor Channel programming for zone No.1

: For sequential channel : For non-sequential channel SET END

Programming mode for zone printing

Zone Nos. 2 to 4

(Note) For RD200 (2 zones only), Screen 2 is

Cursor Channel programming for zone Nos.2 to 4

: For sequential channel : For non-sequential channel

(Note) The programming for zones No. 3 and 4

skipped and Screen 3 appears.

are for RD2800 only.

Show 4 more labels / short text on this page
Zone Compressed/expanded · Number of zones · (2 to 4) · -96-
Page 101 of the Omega RD200 user manual — SHIFT SET END
Page 101 · click to enlarge
Page 101

14. PRINTING FORMAT 14.3 Programming Zone Printing

3 Programming Flow Chart (RD200)

Remedial measure to error display

Programming CH 1 and CH 3 to zone No.1, and CH 2 and CH 4 to zone No. 2

Press any key other than and program again. When it is normal, the next screen appears.

SHIFT

Return to the screen of Step (4)

SHIFT DISP

Ref. 1

For storing parameters into memory, return the screen to Step (4).

SET END

Selection and overlapping of channels

[To be ready for programming]

The channel not selected in any zone is skipped. Also, if the selection of channel is overlapped or the skipped channel is selected, error display appears.

(3 time)→

ºF ENTRY

Checking parameters Reference 1

Press to display parameters.

[Programming channel of 1st area]

The parameters are displayed

sequentially each time is

pressed. Store the parameters into memory. (Programming change mark is printed.) “+” mark is printed at the border of the area. *Fill unused digits with spaces.

[Programming channel of 2nd area]

[To display step (4)]

SHIFT SET END

Border mark Reference 3

Ref. 2

Show 22 more labels / short text on this page
+ · [Checking printing format] · (1) · Standard · (2) · Zone · [Selecting zone printing] · (3) · Various · → · keys · (4) · ∗ · Various · keys · (5) · Storing Reference 2 · ∗ · (6) · + · Stored · -97-
Page 102 of the Omega RD200 user manual — SHIFT SET END
Page 102 · click to enlarge
Page 102

14. PRINTING FORMAT 14.3 Programming Zone Printing

4 Programming Flow Chart (RD2800)

Programming CH 1 and CH 3 to zone No.1, and CH 2 and CH 4 to zone No. 2

If error display appears during programming, press any key other than and program again. When it is normal, the next screen appears. When zones are divided into 3 or 4, program channels up to 3 or 4 zones. The channel not selected in any zone is skipped. Also, if the selection of channel is overlapped or the skipped channel is selected, error display appears. Press to display parameters. The parameters are displayed sequentially each time is pressed. Store the parameters into memory. (Programming change mark is printed.) “+” mark is printed at the border of the area.

Remedial measure to error display

SHIFT DISP

Ref. 1

SHIFT

SET END

[To be ready for programming]

Channels programming for 3rd or 4th zones

° F ENTRY

Remarks 3 Return to the display of Step (4)

ENTRY 2 TAG

For storing the parameter into memory, return the screen to Step (4).

[Programming numbers of area]

Selection and overlapping of channels

Numbers of divided area

[Programming channel of 1st area]

[Programming channel of 2nd area]

Checking parameters Reference 1

Program 3rd area also. Program 4th area with same

procedure.

[To display step (4)]

SHIFT SET END

Border mark Reference 3

Ref. 2

* Fill unused digits with spaces.

Show 27 more labels / short text on this page
+ · [Checking printing format] · (1) · Standard · (2) · Zone scale · (3 times) → · [Selecting zone printing] · (3) · → · (4) · → · Various · keys · (5) · ∗ · → · Various · keys · (6) · ∗ · (7) · Storing Reference 2 · (8) · + · Stored · -98-
Page 103 of the Omega RD200 user manual — NO
Page 103 · click to enlarge
Page 103
Omega RD200 — 7. PROGRAMMING 7.1 Keys and Characters diagram (page 103)
Fig. 217 · 7. PROGRAMMING 7.1 Keys and Characters

This explanation is only for the communications interface (option). Press and simultaneously for 3 seconds or more to display the “communications protocol” programming screen.

This section only explains programming of the communications interface. For details of the communications interface, refer to the separate instruction manual for “Communications Interface”.

SHIFT SPACE COPY(=)

3 sec. or more

SET END

[To be ready for programming]

When EnG is selected and then MODBUS is selected, it is fixed to RTU mode. Select the port type from the followings. (1) : Higher level communications (2) : Engineering Select the protocol from the followings. (1) : Private protocol (2) : MODBUS protocol When [] is selected, the mode selection screen is displayed. Select the mode from the followings. (1) : RTU mode (2) : ASCII mode

Ref. 1

Ref. 2

YES

(ModbUS is selected.)

NO

Ref. 3

Ref. 4

Return to the port selection screen

After “temporarily storage”, the “Port

Ref. 5

selection” screen appears again. For

storing parameters, press . Store the [temporarily stored] parameters into memory. (Programming change mark is printed.)

Show 25 more labels / short text on this page
Communications interface · ( + ) · [Checking port] · Port · % / · [Selecting port] · Port · Reference 1 · Port selection · % / · [Selecting protocol] · Protocol · Reference 2 · Protocol selection · Selected “PriVAtE”? · [Selecting mode] · % / · Reference 3 · Mode selection · Mode · Temporarily stored · Reference 4 · Stored · Storing Reference 5 · -99-
Page 104 of the Omega RD200 user manual — A
Page 104 · click to enlarge
Page 104

15. COMMUNICATIONS INTERFACE 15.2 Programming communications

This explanation is only for the communications interface (option). Press and simultaneously for 3 seconds or more to display the “communications” programming screen.

SHIFT - (-)

2 sec. or more Communications type check

A

Communications type is displayed. The communications type cannot be changed. 01 to 99: (private) protocol 01 to 31: (MODBUS) protocol (Note) “01” is displayed only when “private”

Ref. 7

SHIFT - (-)

3 sec. or more

[Checking communications type]

Ref. 1

is changed to “MODBUS”. 1200, 2400, 4800, 9600,19200 bps Only 9600 and 19200 bps are available with the MODOBUS protocol. (Note) “9600” is displayed only when “private” is changed to “MODBUS”. Characters are displayed with codes.

A

ENTRY SET END

Transmission speed selection range

YES

RS232C?

NO

0 CLEAR 9 LIST

Ref. 2

A

2 sec. or more

length Parity Stop bit

7E1 7 bits Even 1 7E2 7 bits Even 2 701 7 bits Odd 1 702 7 bits Odd 2 8N1 8 bits Non 1 8N2 8 bits Non 2 8E1 8 bits Even 1 8E2 8 bits Even 2 801 8 bits Odd 1 802 8 bits Odd 2 (Note) “8N1” is displayed only when “private” is changed to “MODBUS”. : Checksum ON : Checksum OFF (Note) No display appears for the

Ref. 3

A

2 sec. or more

Ref. 4

A

2 sec. or more

MODBUS protocol. Pressing for each programming executes temporary storing. Pressing for 2 seconds or more executes storing and returns to the operation screen. If the screen returns to the operation screen before pressing , the parameter is returned to the parameter before programming.

Ref. 5

From “ temporarily stored” to “stored”

A

Returning to operation screen

2 sec. or more

Ref. 6

Show 31 more labels / short text on this page
Reference 1 · ( + ) · Address programming range · Reference 2 · → · Reference 3 · to · [Programming address] · Address · Reference 4 · Character selection · Temporarily stored · Code Character · [Selecting transmission speed] · % / · Transmission speed · Temporarily stored · [Selecting characters] · % / · Characters · Reference 5 · Checksum selection range · Temporarily stored · % / · [Selecting checksum] · Reference 6 · Checksum · Temporarily stored · Reference 7 · Stored · -100-
Page 105 of the Omega RD200 user manual — 16. Math Expressions and Totalization
Page 105 · click to enlarge
Page 105

16. Math Expressions and Totalization

Omega RD200 — 16. Math Expressions and Totalization diagram (page 105)
Fig. 218 · 16. Math Expressions and Totalization

This section outlines math expressions and totalization. For programming, read the separate instruction manual for “Math Expressions and Totalizations (Option)”. 1 Math Expressions (1) Measured values are computed and the results are displayed and stored into memory. (2) Math expressions comprise basic calculations, totalization, flow rates, etc., which are depending on the

relevant model code. (3) 18 types of math expressions are provided and they can be programmed to any channel.

Omega RD200 — 16. Math Expressions and Totalization diagram (page 105)
Fig. 219 · 16. Math Expressions and Totalization

(1) None No computation (Display and printing of measured values)

(2) Arithmetical operation 1 +, -. x (Aax + By + Cxy + D)

Rx – Rz/Rs - Rz Rs: Max. range, Rz: Min. range

Computation of measured values by the dry and wet bulbs by using the relative humidity tables

(9) Maximum Maximum value at the programmed “interval period”

(10) Minimum Minimum value at the programmed “interval period”

(11) Average value Average value at the programmed “interval period”

2 Totalizing (1) Measured values and computed results are totalized and the results are displayed and printed. The reset of

the totalization is executed with the programmed interval or a contact signal for the “remote contacts” (option). (2) Totalization is displayed as “int”. This can be programmed to any channel. 3 Instruction manual An instruction manual (INST. No. INE-288) for “Math expression and totalization (option)” is attached separately.

-101-

Page 106 of the Omega RD200 user manual — CH 1
Page 106 · click to enlarge
Page 106

17. Other Options 17.1 Shunt Resistor for Current Input

  • DC current input can be measured by attaching a shunt resistor (option) to the input terminals. 1 Shunt resistor (Option) and Measurement current range
  • A shunt resistor converts the DC current input into a DC voltage. The two types shown in the right table are available.
  • The current measuring ranges are shown in the right table, too. 2 Connection Connect a shunt resister to each channel for the DC current measurement.

Shunt resistor and measuring range

range EZ-RX100 100Ω ±50mA DC

EZ-RX250 250Ω ±20mA DC

Accuracy100 Ω - Rated value ± 0.05%

250Ω – Rated value ±0.1% Resistan

Ex) Connection of a shunt resistor to CH1

Measuring current source Recorder

Only one shunt resistor is to be connected to a channel. 3 [Range/Printing range] and [Scale] 1) Range No. The converted voltage with any shunt resistor is ±5 V DC and the range No. is “07”.

CH 1

100Ω or 250Ω

  • 2) Printing range Program the printing range with the value after being converted into a voltage.
  • Minimum value: Min. input current x shunt resistor value
  • Maximum value: Max. input current x shunt resistor value

Using Range No.

No.

100Ω ±50mA DC ±5VDC 07

250Ω ±20mA DC ±5VDC 07

  • <Input to be printed with the trace printing: 4 to 20mA> Use the shunt resistor of 250 Ω for the maximum measuring current of 20mA.
  • Minimum value: 4 (mA) x 250 ( Ω ) = 1 (V)
  • Maximum value: 20 (mA) x 250 ( Ω ) = 5 (V) <Physical quantity of 4 to 20mA is 0 to 500 litter/m 2 >
  • 3) Scale Program the scale with the physical quantity against the input current.
  • Minimum value: Physical quantity of the minimum input current
  • Maximum value: Physical quantity of the maximum input current

Example of printing range

Example of scale programming

  • Minimum value: 0
  • Maximum value: 500
Show 13 more labels / short text on this page
Code Resistance · Measuring · value* · + · + · Shunt Resistor · Measuring · Converted · Range · ce value · range · voltage · -102-
Page 107 of the Omega RD200 user manual — 1 Transmitter Power Supply
Page 107 · click to enlarge
Page 107

17. Other Options 17.2 Transmitter Power Supply

1 Transmitter Power Supply

This power supply unit, which is designed to install on the rear panel (terminal board) of the instrument, supplies power (24VDC) to a transmitter that transmits the measurement input signals to your recorder.

[Model] RZ-TPS01 [Instruction manual] INST. No. INE-277

-103-

Page 108 of the Omega RD200 user manual — 2 Preparation
Page 108 · click to enlarge
Page 108

18. ADJUSTMENT 18.1 Adjustment of Measured Values

  • Adjustment

Adjustment comprises four kinds shown below. (1), (3) and (4) have already been adjusted. However, it is recommended for maintaining the measuring and printing accuracy to adjust them once a year.

Omega RD200 — 18. ADJUSTMENT 18.1 Adjustment of Measured Values diagram (page 108)
Fig. 220 · 18. ADJUSTMENT 18.1 Adjustment of Measured Values

(1) Measured values adjustment Adjustment to keep the measured value against input within the accuracy ratings. Sec.18.1

measured value Programming to shift a measured value Sec.18.2

(3) Adjustment of printing position Adjustment to set the printing range to the zero and span lines on the chart Sec.18.3

Adjustment of the gaps between pens for the time-axis synchronization Sec.18.4

1 Adjustment of measured values

Omega RD200 — 1 Adjustment of measured values diagram (page 108)
Fig. 221 · 1 Adjustment of measured values

It is recommended for maintaining measuring and printing accuracy to adjust them once a year. (1) Execute the adjustment for each channel. (2) Execute the adjustment under the reference

Ambient temperature 23 ±2°C Ambient humidity 55 ±10%RH Power voltage 100 ±1VAC Power supply

condition. (See the right table.)

2 Preparation

1. Preparation of tools
Omega RD200 — 2 Preparation diagram (page 108)
Fig. 222 · 2 Preparation

DC standard voltage/current generator ○ ○ Accuracy: Should be better than ±0.05%.

compensator ○ 0°C ±0.2°C or less

Thermocouple for test ○ Same type of thermocouple as input type

Standard variable resistor ○ Accuracy: Should be better than ±0.05%.

3-core copper wire ○ Same resistance for each 3 cores

2. Connection Connections depend upon the input types. See the next page.
3. Before starting adjustment (1) After the connection, mount the terminal board cover and turn on the power supply. (2) Before starting adjustments, warm up your recorder for at least 30 minutes until it becomes stable. (It is

recommended to warm it up for at least one hour if possible.)

Omega RD200 — 2 Preparation diagram (page 108)
Fig. 223 · 2 Preparation

Check and adjustments of the measured values need meticulous work in addition to the standard tools and reference conditions. If you need checks or adjustments of the measured values, please consult your nearest CHINO’s agent.

-104-

Page 109 of the Omega RD200 user manual — (A)
Page 109 · click to enlarge
Page 109

18. ADJUSTMENT 18.1 Adjustment of Measured Values

3 Connections

Connections depend upon the input types. Connect the standard tools to the input terminals to be adjusted.

Make sure to connect after turning off the power supply. Caution

To prevent any injury caused by electric shock, be sure to turn off the power supply before connections. (1) In case of thermocouple input

1 2 3 4

Thermocouple wire * Copper lead wire

Omega RD200 — 3 Connections diagram (page 109)
Fig. 224 · 3 Connections
Omega RD200 — 3 Connections diagram (page 109)
Fig. 225 · 3 Connections

Ice + distilled water

The electromotive force of the thermocouple input is reduced by an amount proportional to the temperature at the terminals. The recorder itself compensates (reference junction compensation) for this reduction. The adjustment is executed by inputs based on the reference electromotive force (0°C reference). In case the “RJ selection” is set to “1” (Enable) in the “Range/Printing range” programming, an amount equivalent to the reference junction compensation has to be subtracted using a reference junction compensator.

(2) DC voltage input (3) Resistance thermometer input

1 2 3 4

1 2 3 4

(A)

(B)

(B)

Show 21 more labels / short text on this page
(+) · ( - ) · Test tube · Input terminals · Silicon oil · + - · DC standard voltage · Vacuum bottle · generator · Reference junction · compensator · (+) · ( - ) · Standard variable · + - · Input terminals · resistor · Input terminals · DC standard voltage · generator · -105-
Page 110 of the Omega RD200 user manual — SHIFT SET END
Page 110 · click to enlarge
Page 110

18. ADJUSTMENT 18.1 Adjustment of Measured Values

4 Programming Flow Chart

Adjustment of CH 01 (Printing range: -50 to 150)

Computed correction data are canceled if the screen is returned to the operation screen in the procedures before [Storing]. When a skipped channel is programmed, operation is not accepted. Change the connection to other input terminals to be adjusted and repeat the same procedure from the programming channel screen (indicated by a dotted line). In zero adjustment or span adjustment screen, when and are pressed simultaneously and then is pressed, the correction data of its channel are initialized. : Memory clear : Adjustment : Hardware check : Measured value adjustment : Shift programming for

By returning to the operation screen

2 sec. or more

SHIFT 6 A -Z

3 sec. or more

Ref. 1

Skipped channel does not accept any [ENTRY] key

° C ° F ENTRY

° F ° C ENTRY

Ref. 2

1 CHART

Initialization of correction data

[To zero adjustment display]

SHIFT 0 CLEAR

[Zero adjustment] Apply an input equivalent to the minimum value with a standard tool.

(Computation of zero correction data)

Engineering mode types Reference 1

[To span adjustment display]

[Span adjustment] Apply an input equivalent to the maximum value with a standard tool.

(Computation of span correction data)

[To programming channel screen]

measured value : Printing position adjustment : Time-axis adjustment for pens Store the computed correction data into memory.

SHIFT SET END

Ref. 3

Show 23 more labels / short text on this page
( + ) · [Engineering mode selection] · Cursor Adjustment · → · [Selecting adjustment] · → · Other channels adjustment · [Selecting measured value] · Measured value Cursor · [Programming channel] · Channel Cursor · (Zero) · Minimum value · (Span) · Maximum value · Reference 2 · Adjustment types · Channel Cursor · + · Storing · Stored · Reference 3 · -106-
Page 111 of the Omega RD200 user manual — 0 CLEAR SHIFT
Page 111 · click to enlarge
Page 111

18. ADJUSTMENT 18.2 Shift Programming of Measured Value

This programming is for slight-shifting a measured value. After programming, measured values are shifted by the programmed amount. 1 Shift Programming

Example of shift programming

Measured value ….850.3

(1) This programming is necessary for each channel. (2) The cursor shifts to the least significant digit. Execute this

Programming …… 850.0

programming with or . 2 Programming Flow Chart

Shift value = 850.0-850.3 = -0.3 Before starting this programming, wait for at least 30 minutes after turning on the power supply. Programmed shifting data is canceled if the screen is returned to the operation screen in the procedures before [Storing]. When a skipped channel is programmed, operation is not accepted. Change the connection to other input terminals to be adjusted and repeat the same procedure from the programming channel display (indicated by a dotted line). In zero adjustment or span adjustment screen, when and are pressed simultaneously and then is pressed, the correction data of its channel are initialized. : Memory clear : Adjustment : Hardware check : Measured value adjustment : Shift programming for

Changing measured value of CH 1 from 850.3 to 850.0

Omega RD200 — programming with or . 2 Programming Flow Chart diagram (page 111)
Fig. 226 · programming with or . 2 Programming Flow Chart

By returning to the operation screen

Skipped channel does not accept any [ENTRY]

Initialization of shift data (0)

0 CLEAR SHIFT

measured value : Printing position adjustment : Time-axis adjustment for pens Store the programmed shifting data into memory.

Show 11 more labels / short text on this page
% · % / · Before programming · Other channels adjustment · Reference 1 · Engineering mode types · Adjustment types · Reference 2 · Reference 3 · Storing · -107-
Page 112 of the Omega RD200 user manual — 18. ADJUSTMENT 18.3 Adjustment of Printing Position
Page 112 · click to enlarge
Page 112

18. ADJUSTMENT 18.3 Adjustment of Printing Position

This is the adjustment for the printing position of the cartridge pen for use in trace printing. It is recommended to adjust it once a year for maintaining the printing accuracy. ■ Adjustment Flow Chart

Omega RD200 — 18. ADJUSTMENT 18.3 Adjustment of Printing Position diagram (page 112)
Fig. 227 · 18. ADJUSTMENT 18.3 Adjustment of Printing Position
Omega RD200 — 18. ADJUSTMENT 18.3 Adjustment of Printing Position diagram (page 112)
Fig. 228 · 18. ADJUSTMENT 18.3 Adjustment of Printing Position

Pressing once moves 0.1 mm to right. Pressing once moves 0.1 mm to left. Press when the printing position matches to zero or span of the chart. Pens stand by near the scale center. Computed correction data are canceled if the screen is returned to the operation screen in the procedures before [Storing]. ∗During zero and span adjustments, operation is not accepted. : Memory clear : Adjustment : Hardware check : Measured value adjustment : Shift programming for

Other pens than the pen being adjusted

By returning to the operation screen

To match the printing position to “Zero”, press few times.

Omega RD200 — 18. ADJUSTMENT 18.3 Adjustment of Printing Position diagram (page 112)
Fig. 229 · 18. ADJUSTMENT 18.3 Adjustment of Printing Position

measured value : Printing position adjustment : Time-axis adjustment for pens 1: 1st pen 2: 2nd pen (2-pen type to 4 pen type) 3: 3rd pen (3-pen type, 4-pen type) 4: 4th pen (4-pen type only) Store the computed correction data into memory.

To match the printing position to “span”, press few times.

Omega RD200 — 18. ADJUSTMENT 18.3 Adjustment of Printing Position diagram (page 112)
Fig. 230 · 18. ADJUSTMENT 18.3 Adjustment of Printing Position

Reference 1 · Engineering mode types · -108-

Page 113 of the Omega RD200 user manual — YES
Page 113 · click to enlarge
Page 113

18. ADJUSTMENT 18.4 Time-axis Adjustment of Pens

When the time-axis synchronization (POC) is programmed to “on”, the gaps between the pens may change over time, resulting in error between their time-axis. This adjustment is for correction of these changes. It is recommended to adjust them once a year for maintaining the time-axis synchronization. ■ Adjustment Flow Chart

Pens stand by at 0% position of scale. : Memory clear : Adjustment : Hardware check : Measured value adjustment : Shift programming for

prints a straight line.

2 sec. or more

(3) When these two lines match, press . ENTRY

SHIFT 6 A - Z

3 sec. or more

Ref. 1

(2) Initial position of the pen adjusted

Other pens than the pen being adjusted

° C ° F ENTRY

° C ° F ENTRY

Engineering mode types Reference 1

Ref. 2

° C ° F

Adjustment types Reference 2

[Selecting an adjusting pen]

Ref. 3

measured valuePrinting position adjustment : Time-axis adjustment for pens

(1) (2) (3) (4)

When (1) the reference pen prints a straight line on 0% to 100% of the chart and (2) Press , the chart is fed and then the printing of the adjusting pen coincides with the straight line (Note) Note) If you press while it does not coincide

adjusted (1) 1 P Plotter pen 1st pen

(2) 2 P 1st pen 2nd pen

with the straight line, re-execute the procedure from the beginning of [Selecting an adjusting pen]. (Computation of correction data)

(3) 3 P 1st pen 3rdt pen

(4) 4 P 1st pen 4th pen

Store the computed correction data into memory.

YES

NO SHIFT SET END

Ref. 4

Show 23 more labels / short text on this page
Adjustment procedure · (1) Reference pen · ( + ) · [Engineering mode selection] · Cursor Adjustment · → · [Selecting adjustment] · → · [Selecting time-axis] · Time axis · (Starting adjustment ) · Reference 3 · Adjusting pen selection · / · Display Reference · Pen to be · pen · Temporarily storing · Storing Reference 4 · Programming other channel? · + · Stored · -109-
Page 114 of the Omega RD200 user manual — 1 Check Flow Chart
Page 114 · click to enlarge
Page 114

19. HARDWARE CHECK 19.1 ROM Version Check

  • Hardware check

Hardware check comprises the following seven items. Items (5) to (7) only apply to corresponding options added.

Omega RD200 — 19. HARDWARE CHECK 19.1 ROM Version Check diagram (page 114)
Fig. 231 · 19. HARDWARE CHECK 19.1 ROM Version Check

(1) ROM version check Checking of versions for ROM and linearization Sec. 19.1

(2) Printer test Checking of printing with the cartridge pens and the plotter pen Sec. 19.2

(3) Display test Checking of LCD on the display Sec. 19.3

(4) Measuring input check Checking of real data (A/D converted value) of each channel Sec. 19.4

(5) Alarm output check Output of on or off signal from terminals Sec. 19.5 (6) Remote contacts input

check Checking of input condition (open or short) at terminals Sec. 19.6

1 Check Flow Chart

: Memory clear : Adjustment : Hardware check (1): ROM version check (2): Printer check (3): Display check (4): Measuring input check Press or to check the versions of ROM or linearization.

Omega RD200 — 1 Check Flow Chart diagram (page 114)
Fig. 232 · 1 Check Flow Chart

-110-

Page 115 of the Omega RD200 user manual — REC ON/OFF ENTRY
Page 115 · click to enlarge
Page 115

19. HARDWARE CHECK 19.2 Printer Check

This is for checking of trace printing with the cartridge pens and digital data printing with the plotter pen. 1 Example of Printing Check – When the time-axis synchronization (POC) is “off” in 3-pen type

2 Check Flow Chart

Omega RD200 — 2 Check Flow Chart diagram (page 115)
Fig. 233 · 2 Check Flow Chart

1. Digital data printing (Plotter pen) Two lines are printed. The number of characters per line is 40 with AL and 72 with the AH (repetition of A to Z, 1 to 9 and 0). 2. Trace printing (Cartridge pens) Each pen prints a sine curve. (Note) Even when the time-axis synchronization is on, printing s executed in condition of the time-axis synchronization “off” during printing check. * is not accepted during printing check. For cancellation of printing check, press and then press . By pressing for 2 seconds or more, the screen returns to the operation screen, but the printing disables. For changing the printing to enable, press and then press again. See Reference 1 and Reference 2 in section 19.1.

2 sec. or more

SHIFT 6 A - Z

3 sec. or more

° C ° F

° C ° F

Cancellation of printing check

REC ON/OFF ENTRY

REC ON/OFF ENTRY

Digital data printing starts first.

REC ON/OFF ENTRY

Types of engineering mode

Show 16 more labels / short text on this page
∗ · Check printing details · ( + ) · [Engineering mode selection] · Reference · Cursor Check · [Selecting check] · [Selecting check items] · [Selecting printer] · [Displaying check] · [Executing check] · → · Reference · and check 1 · Stop checking operation · -111-
Page 116 of the Omega RD200 user manual — ° C ° F
Page 116 · click to enlarge
Page 116

19. HARDWARE CHECK 19.3 Display Check

This is for checking of 16- or 7-segment LCD, status lamps and bargraphs. 1 Display …. (Note) The following is for RD200. It is same for RD2800.

2 Check Flow Chart

Omega RD200 — 2 Check Flow Chart diagram (page 116)
Fig. 234 · 2 Check Flow Chart

(1) All segments and lamps light. (2) 16-segment LCD and underbar (3) 7-segment LCD and underbars

Order of check lighting

2 sec. or more

SHIFT 6 A - Z

(From left to right) (4) Programming lamp, memory colon

3 sec. or more

Ref. 1

and status lamps (all) (5) Bargraph on the upper row

(5-segment at a time, from left to right) (6) Bargraph on the second row

° C ° F ENTRY

(Same as the upper row) → On the third row → On the fourth row (7) All segments and lamps light out. For cancellation of display check, press for 2 seconds or more.. The screen returns to the operation screen. : Memory clear : Adjustment : Hardware check (1): ROM version check (2): Printer check (3): Display check (4): Measuring input check

° C ° F

Ref. 2

Cancellation of display check

(1) (2) (3) (4)

ENTRY Engineering mode types Reference 1

All segments and lamps light first.

Types of check 1 Reference 2

Show 9 more labels / short text on this page
( + ) · [Engineering mode selection] · Cursor Check · → · [Selecting check] · [Selecting a display] · [Displaying check] · [Executing check] · -112-
Page 117 of the Omega RD200 user manual — RJ
Page 117 · click to enlarge
Page 117

19. HARDWARE CHECK 19.4 Measuring Input Check

The measuring input can be checked by counts after A/D conversion. ■ Check Flow Chart

: Memory clear : Adjustment : Hardware check (1) ROM version check (2) Printer check (3) Display check (4) Measuring input check CH1 to CH4 show the counts after A/D conversion of inputs to channels 1 to 4. For RD2800, the RJ (reference junction compensation) count is displayed next to CH4. RJ is the counts after A/D conversion of the reference junction compensation (RJ). RD2800 has no Count 2 display as RJ is displayed in Count 1 display. RJ is displayed only when there is a channel, of which RJ is programmed to “1: enable” in the range programming. When all channels are programmed to “0: disable”, (-) is displayed.

Engineering mode types Reference 1

2 sec. or more

SHIFT 6 A – Z

3 sec. or more

Ref. 1

Types of check 1 Reference 2

° C ° F ENTRY

° C ° F

Count 1 display Reference 3

Ref. 2

(1) (2) (3) (4)

Count 2 display Reference 4

[Displaying count1]

Ref. 3

Ch1 Ch2 Ch3 Ch4

Ref. 4

RJ

Show 10 more labels / short text on this page
( + ) · [Engineering mode selection] · Cursor Check · → · [Selecting check ] · [Selecting measuring input] · [Displaying check] · % / · [Displaying count 2] · -113-
Page 118 of the Omega RD200 user manual — 19. HARDWARE CHECK 19.5 Alarm Output Check
Page 118 · click to enlarge
Page 118

19. HARDWARE CHECK 19.5 Alarm Output Check

This check is only available for the alarm output (option). It checks the relay drive circuits and relays by outputting the shorted (on) or open (off) signals at the specified alarm output terminals (relay Nos.). ■ Check Flow Chart

* N.C terminals are provided for the mechanical relay “c” contact output.

Omega RD200 — 19. HARDWARE CHECK 19.5 Alarm Output Check diagram (page 118)
Fig. 235 · 19. HARDWARE CHECK 19.5 Alarm Output Check
Omega RD200 — 19. HARDWARE CHECK 19.5 Alarm Output Check diagram (page 118)
Fig. 236 · 19. HARDWARE CHECK 19.5 Alarm Output Check

(Common for “a” and “c” contacts) On 0.1Ω ore less

Press or to switch Check 1 and 2. (1): Alarm output check (2): Remote contacts input check (3): Communications interface check The programming range of relay Nos. are 1 to 6 for AL and 01 to 12 for AH. The selection of output (on or off) is effective at the moment is pressed. The previous output status is kept until then.

-114-

Page 119 of the Omega RD200 user manual — 19. HARDWARE CHECK 19.6 Remote Contacts Input Check
Page 119 · click to enlarge
Page 119

19. HARDWARE CHECK 19.6 Remote Contacts Input Check

This check is only available for the remote contacts (option). It checks the input signals (shorted or open) at the remote contacts terminals. ■ Check Flow Chart

Eleven kinds of operations are available by the remote contacts signals. The operation allocated to each terminal (EX1 to EX4) can be checked. ⇒ See Section 13.2. Press or to switch Check 1 and 2. (1): Alarm output check (2): Remote contacts input check (3): Communications interface check The input signals at EX1 to EX4 are:

Omega RD200 — 19. HARDWARE CHECK 19.6 Remote Contacts Input Check diagram (page 119)
Fig. 237 · 19. HARDWARE CHECK 19.6 Remote Contacts Input Check

-115-

Page 120 of the Omega RD200 user manual — RS-232C RS-422A RS-485
Page 120 · click to enlarge
Page 120

19. HARDWARE CHECK 19.7 Communications IF Check

This check is only available for the communications interface (option). It checks the communications function by displaying the signal received after transmitting it. 1 Connections

Except for RS-485, short the transmission and reception terminals.

RS-232C RS-422A RS-485

SA RA SG SD RD SG

SDA SDB RDA RDB SG

Short SD and RD terminals. Short SDA and RDA terminals. Short SDB and RDB terminals.

Open circuit. (No need to short.)

Show 4 more labels / short text on this page
Shorted · Shorted · Shorted · -116-
Page 121 of the Omega RD200 user manual — 2 Check Flow Chart
Page 121 · click to enlarge
Page 121

19. HARDWARE CHECK 19.7 Communications IF Check

2 Check Flow Chart

Press or to switch Check 1 and 2. (1): Alarm output check (2): Remote contacts input check (3): Communications interface check 1. Communications type Communications type installed is displayed.

Omega RD200 — 2 Check Flow Chart diagram (page 121)
Fig. 238 · 2 Check Flow Chart
  • 2. Communications status display
  • Normal condition: Displays one digit by one digit as 0 → 1 → 2 …9, and after 9, repeat from 0 again.
  • Abnormal condition: One of 0 to 9 blinks. In this case, communications circuit is abnormal.

-117-

Page 122 of the Omega RD200 user manual — 20. MEMORY CLEAR
Page 122 · click to enlarge
Page 122

20. MEMORY CLEAR

This is for initializing parameters and correction data to the default values. 1 Initialization items

Omega RD200 — 20. MEMORY CLEAR diagram (page 122)
Fig. 239

Parameters Initializes all parameters excluding [time] to the default values. Refer to Section 7.3 (List of programming items) for the default values.

Correction data Initializes all correction data to the default values.

The time is preprogrammed to Japanese time at shipment. When initializing the time, it becomes 00:00 hours on January 1, 2000. If the operation screen is not shown due to a failure of the LSI for time, clear the programmed time. 2 Operation Flow Chart

Omega RD200 — 20. MEMORY CLEAR diagram (page 122)
Fig. 240

(3): Clock Press to initialize the selected item. After initialization, the screen returns to the [Displaying items] screen.

-118-

Page 123 of the Omega RD200 user manual — 2 Operation check
Page 123 · click to enlarge
Page 123

21. MAINTENANCE 21.1 Daily Inspection

Check the residual quantity of chart paper, displaying/printing conditions, etc. on a daily basis in order to use your recorder under good conditions. 1 Consumable parts check

Omega RD200 — 21. MAINTENANCE 21.1 Daily Inspection diagram (page 123)
Fig. 241 · 21. MAINTENANCE 21.1 Daily Inspection
  • Check the residual quantity of chart. When the residual quantity becomes less, a message “Prepare a new chart” or “New Paper Required” appears with red ink on the left of the chart paper.
  • Loading chart paper See Section 5.1.
  • Ordering chart paper See Section 1.2.
2. Plotter pen If the text on the digital data printing becomes unclear or pale, replace the plotter pen with a new one. See Section 5.2.
3. Cartridge pen If the line on the trace printing becomes unclear or pale, replace the cartridge pen with a new one. See section 5.2.

2 Operation check

Omega RD200 — 2 Operation check diagram (page 123)
Fig. 242 · 2 Operation check

Check that there is no paper jam or dislocation from the sprocket. (2) Text and trend line printing condition

Check that no disorder or abnormal condition is found in the text and trend line printing.

2. Display conditions Check that no abnormalities are found in the displays of measured values, status and bargraph.
3. Others (1) Check that the measured values are normal. (2) Check if any noise or strange odors are generated.

-119-

Page 124 of the Omega RD200 user manual — 1 Main Shaft of Plotter Pen
Page 124 · click to enlarge
Page 124

21. MAINTENANCE 21.2 Cleaning and Lubrication

Clean the main shaft of each pen once a year in order to maintain a satisfactory printing performance.

Plotter pen Cartridge pen (RD200) Cartridge pen (RD2800)

Omega RD200 — 21. MAINTENANCE 21.2 Cleaning and Lubrication diagram (page 124)
Fig. 243 · 21. MAINTENANCE 21.2 Cleaning and Lubrication
Omega RD200 — 21. MAINTENANCE 21.2 Cleaning and Lubrication diagram (page 124)
Fig. 244 · 21. MAINTENANCE 21.2 Cleaning and Lubrication
Omega RD200 — 21. MAINTENANCE 21.2 Cleaning and Lubrication diagram (page 124)
Fig. 245 · 21. MAINTENANCE 21.2 Cleaning and Lubrication

1 Main Shaft of Plotter Pen

2 Main Shaft of Cartridge Pen

(1) Turn off printing. Pull out the chart cassette and

(1) Turn off printing. Pull out the chart cassette and

open the display section. (2) Wipe off dirt from the main shaft with a cotton

open the display section. (2) Wipe off dirt from the main shaft with a cotton

bud or similar tool. (Dirt can be wiped off more easily with a cotton bud soaked in alcohol.) (3) Apply one or two drops of the provided

bud or similar tool. (3) Apply one or two drops of the provided

lubricating oil to the main shaft.

lubricating oil to the main shaft.

(1) Do not move the plotter pen or cartridge pen by hand. (2) Use the provided lubricating oil. Do not use any other oil. 3 Cleaning the Door

The door is made of plastic*. Clean it with a dry soft cloth or with a soft cloth moistened with lukewarm water or a neutral detergent. * Aluminum die-cast door is available an option. (RD200)

Do not use any chemicals

Do not use thinner, benzene or other chemicals that may damage the plastic components. These may cause deformation or breakage to the door.

Show 4 more labels / short text on this page
Main shaft · Main shaft · Main shaft · -120-
Page 125 of the Omega RD200 user manual — 23 ± 2 ° C. 2 Preparation
Page 125 · click to enlarge
Page 125

21. MAINTENANCE 21.3 Measured Values Check

It is recommended for check measured values once a year for maintaining the measuring and printing accuracy. 1 Channels to be checked

(Note) Add ±0.01% per 1°C for a case other than

Check the measured value for each channel. Different errors may occur for different channels even in the same range.

Items Reference conditions Ambient temperature 23 ±2°C

Ambient humidity 55 ±10%RH Power voltage 100 V AC ±1% Power supply

To check a channel for which a math-function or totalization (optional) has been programmed, program the math expression to “” before starting the checking.

frequency 50 or 60 Hz ±0.5 Hz

23 ± 2 ° C. 2 Preparation

1. Preparation of tools

DC standard voltage/current generator { { Accuracy: Should be better than ±0.05%. Reference junction

compensator { 0°C ±0.2°C

Thermocouple for test { Same type of thermocouple as input type

Standard variable resistor { Accuracy: Should be better than ±0.05%.

3-core copper wire { Three copper wires shall have the same resistance values. 2) Connection Connections depend upon the input types. See the next page.

3. Before starting adjustments (1) Mount the terminal board cover and turn on the power supply. (2) Before starting adjustments, warm up your recorder for at least 30 minutes until it becomes stable. (It is

recommended to warm up for at least one hour if possible.)

Check and adjustments of the measured values need meticulous work in addition to the standard tools and reference conditions. If you need checks or adjustments of the measured values, please consult your nearest CHINO’s agent.

Show 9 more labels / short text on this page
Reference conditions · Totalization channel · Input types Tools · DC voltage Thermo-coupl · Resistance · e · thermometer · Checking · -121-
Page 126 of the Omega RD200 user manual — T
Page 126 · click to enlarge
Page 126

21. MAINTENANCE 21.3 Measured Values Check

3 Connections

Connections depend upon the input types. Connect both standard and other tools to the measuring input terminals to be adjusted.

Turn off the power source before starting connections Caution

In order to prevent electric shock, turn off the power source before starting connections.

(1) In case of thermocouple input

*Thermocouple of the same type as the input type

1 2 3 4

Thermocouple wire * Copper lead wire

Ice + distilled water

The electromotive force of the thermocouple input is reduced by an amount proportional to the temperature at the terminals. The recorder itself compensates (reference junction compensation) for this reduction. The adjustment is executed by inputs based on the reference electromotive force (0°C reference). In case the “RJ selection” is set to “1” (Enable) in the “Range/Printing range” programming, an amount equivalent to the reference junction compensation has to be subtracted using a reference junction compensator.

(2) DC voltage input (3) Resistance thermometer input

1 2 3 4

1 2 3 4

(A)

(B)

(B)

4 Checking Method

Input the printing range (T) of 0%, 50% and 100% from a standard tool and obtain an error (e) by reading the measured values (M).

M – T

e (%) = X 100

T

Show 21 more labels / short text on this page
(+) · ( - ) · Input terminals · Test tube · Silicon oil · + - · DC standard voltage · Vacuum bottle · generator · Reference junction · compensator · (+) · (-) · + - · Input terminals · Standard variable · resistor Input terminals · Standard variable · resistor · Error computation · -122-
Page 127 of the Omega RD200 user manual — 2 Measured Value Abnormal
Page 127 · click to enlarge
Page 127

21. MAINTENANCE 21.4 Troubleshooting

Troubleshooting methods are shown classified by symptoms. Read the item that corresponds to the symptom.

Repair and modifications Caution

Never repair or modify the instrument by replacing assembled component units or parts, otherwise correct repair or modifications cannot be executed and also electric shock or damage of your recorder may occur. 1 Not Working

Check Causes and remedial measures 1) Check if power is supplied to the power terminals. Turn on the external power supply source.

2. Check if the power supply is as specified. Supply the specified power (100 to 240V AC and 50/60 Hz). 3) Check if the connections to the power terminals are correct. Connect the cable to the power terminals (L, N) correctly.
4. Try turning off or on the external power supply source.
5. Initialize the parameters and check if your recorder recovers to normal conditions. See Section 20. (Caution) All parameters become the default values. Program them again.

2 Measured Value Abnormal

Symptoms Causes and remedial measures

  • Check if the measuring terminals are loose.
  • Check if the input signal is unstable.
  • Check if the thermocouple is connected with another instrument in parallel and the burnout function is programmed to [Enable].
1. Measured values are unstable.
2. Check if the measured value is displayed as shown below.
  • Check if the input terminals are connected correctly.
  • Check if the input terminals are loose.
  • Check if the input signal wires are disconnected.
  • Check if the input signal exceeds the measuring range.
  • Check if the selection of ºC/ºF computation is correct.
  • Check if the input signal is correct.
  • Check if an extension wire is connected to the input terminal. (Thermocouple input type only)
  • Check the scale and adjust the measured values if any error is detected. (See Section 18.1.)
3. An error occurs.
  • Check if the RJ selection is 0 (disable) in [Range/Printing range] programming. (This is negligible if reference junction compensation is executed externally.)
  • Check if the terminal cover is mounted.
4. Influences by ambient temperature (Thermocouple input type only)

-123-

Page 128 of the Omega RD200 user manual — SHIFT SHIFT
Page 128 · click to enlarge
Page 128

21. MAINTENANCE 21.4 Troubleshooting

3 Display Abnormal

Symptoms Causes and remedial measures

1. Measured value blinks. Measured value of the channel in which an alarm activates. See Section 10.3. 2) Measured value is blank. No measured value is displayed for the skipped channel. 3) Year, month, day, hour and minute are displayed. This is one of the operation screens (Clock display). See Section 6.4.
4. “ ” is displayed at the left end. This is one of the operation screens (Alarm activation screen). See Section 6.4. 4 Printing Abnormal

Symptom Causes and remedial measures

  • 1) No printing is executed.
  • Check if the printer is turned off. See Section 6.2.
  • Check if the cassette ribbon is mounted normally.
2. Digital printing is not executed. All digital printings are not executed if the chart speed is programmed to be faster than 150mm/h. See Section 10.2.
3. Printing color is pale. Replace the cartridge pen or the plotter pen.
  • Check if the chart paper is unlatched from the sprocket.
  • Check if the chart paper cassette is securely inserted into the internal unit.
  • Check if the chart paper has been shuffled before loading it. 5 Other Troubles
4. Chart paper feed is abnormal.

Symptoms Causes and remedial measures

1. Keys are not acceptable.
2. Parameters cannot be changed. Keys are locked. Unlock them. See Section 11.12.

To store parameters to be programmed for each channel, press

and simultaneously. If not, parameters are not stored into memory. See Section 8.2 and other sections for programming.

3. Prarameters remain unchanged after changing them.

SHIFT SHIFT

If the above troubleshooting cannot solve your problem, certain parts may be defective. Check the following items and contact your CHINO’s sales agent. (1) Model (2) Serial No. (3) Details of trouble (4) Other symptoms found

Request · -124-

Page 129 of the Omega RD200 user manual — 2 Recommended intervals of parts replacement
Page 129 · click to enlarge
Page 129

21. MAINTENANCE 21.5 Recommended Parts Replacement Intervals

It is recommended to replace parts periodically as a preventive measure to enable the use of your recorder under ideal conditions over long periods of time.

Omega RD200 — 21. MAINTENANCE 21.5 Recommended Parts Replacement Intervals diagram (page 129)
Fig. 246 · 21. MAINTENANCE 21.5 Recommended Parts Replacement I…

Do not replace any parts other than consumable chart paper and pens, otherwise your recorder cannot be recovered correctly and a dangerous accident may occur. Consult CHINO’s sales agent for replacing parts. 1 Operation conditions

The recommended parts replacement intervals apply when your recorder is used under the following conditions. The replacement intervals could be shortened if the ambient conditions are worse than the followings.

Omega RD200 — 21. MAINTENANCE 21.5 Recommended Parts Replacement Intervals diagram (page 129)
Fig. 247 · 21. MAINTENANCE 21.5 Recommended Parts Replacement I…
Omega RD200 — 21. MAINTENANCE 21.5 Recommended Parts Replacement Intervals diagram (page 129)
Fig. 248 · 21. MAINTENANCE 21.5 Recommended Parts Replacement I…

Temperature 20 to 25°C Humidity 20 to 80% RH Operation hours 8 hours/day

(1) A place free from dust, moisture or soot (2) A place free from vibrations or shock. (3) A place where no unfavorable conditions

2 Recommended intervals of parts replacement

1. Parts to be mounted (Consumable parts)
Omega RD200 — 2 Recommended intervals of parts replacement diagram (page 129)
Fig. 249 · 2 Recommended intervals of parts replacement

(RD200) 16 m (Option) 30 days In case of continuous use with a chart speed of 20 mm/h

Chart paper (RD2800) 30 days In case of continuous use with a chart speed of 25 mm/h

Cartridge pen 2km Could be shorter than 2 km depending on chart paper speed, pen speed and ambient temperature/humidity.

Could be less than 100,000 characters depending on chart paper speed and ambient temperature/humidity.

2. Component parts
Omega RD200 — 2 Recommended intervals of parts replacement diagram (page 129)
Fig. 250 · 2 Recommended intervals of parts replacement

Pen servo (for cartridge pen) 4 to 6 years

Pen servo (for plotter pen) 4 to 6 years

Motor (for chart paper) 4 to 6 years Motor (for plotter pen) 4 to 6 years Motor (for cartridge pen) 4 to 6 years

Power supply unit 5 years Under an ambient temperature of 25°C Display unit 4 to 6 years Sheet switch (key) 4 to 6 years

70,000 times Resistive load (rated contact capacity or less) Mechanical relay for alarm output 20,000 times Inductive load (rated contact capacity or less) EEPROM 7 years Able to re-write about 100,000 times or less

-125-

Page 130 of the Omega RD200 user manual — 2 Printing Printing accuracy: Measuring accuracy + ± .0.3% of printing
Page 130 · click to enlarge
Page 130

22. SPECIFICATIONS

1 Input Specifications

±10 V DC or less (range of ±2 V or less) ±60 V DC or less (range of ±5 V or less)

Number of measuring points: 1, 2, 3 and 4 points Input signals: DC voltage (mV) …±13.8, ±27.6, ±69.0, ±200

±500 DC voltage (V)…±2, ±5, ±10, ±20, ±50 (Built-

Resistance thermometer ±6 V DC or less

Input correctionShift programming for each channel Input filter: Program primary delay filter

in voltage divider for ±5 V or more) DC current…Available by adding shunt

resistor (option) Thermocouple…B, R, S, K, E, J, T, N,

1 to 10 sec. and “0” (none) for each channel. Maximum common mode voltage: 30VAC Common mode rejection ratio: 140 dB or more (50/60 Hz) Series mode rejection ratio: 50 dB or more (50/60 Hz) Terminal board: Detachable and removable for wiring

Platinel II, U, L Resistance thermometer…Pt100 (1),

2 Printing Printing accuracy: Measuring accuracy + ± .0.3% of printing

Pt100 (2), JPt100, Pt50, Pt-Co Range: Program range number and printing range for each

channel with key operation. (See Section 8.2 for range No. list) Scale: Program minimum and maximum values with key

range (under the reference operation

operation. Programming range…-9999 to 99999 Decimal point … Program arbitrary Measuring range: Refer to the list of range No. in Section

Printing dead and0.2% of printing range

Printing system:

Trace printing… Disposable cartridge pen

Digital data printing …. Plotter pen

8-2. Minimum printing rangeRefer to the following table
Step responseRD200 … 1 sec. or less (90% response)

DC voltage 1/5 or higher than the measuring range.

RD2800 … 1.5 sec. or less (90% response)

Print colorTrace printing …. 1 st pen (Red), 2 nd pen (Green),

2/5 or higher than the measuring range when converted into electromotive force. (In the case of range Nos. with minus measuring ranges, a temperature of 0°C or higher is specified as the lower-limit measuring value.)

3rd pen (Blue), 4th pen (Brown)

Digital data printing … Purple

<<Digital printing>> cPeriodic data printing

: Printing of time, channel No., measured

values and engineering unit with programmed interval dDigital data printing

Pt100 100°C span or more Resistance thermometer Pt50 200°C span or more Accuracy rating: Refer to item 6 (at reference operating condition). Temperature drift: ±0.01% full scale/°C (converted into

: Printing of time, channel No., measured

values and engineering unit with trace printing on demand eYear, Month,

: Printing of year, month, day and time when

reference range) Measuring cycleAbout 100 msec (CH1 to CH4 ) Reference junction compensation : Inside (enable) or

turning on the power. Printing time line and time every on the hour. Printing year, moth and day at 00:00 and every specific interval fChannel No.,

day, time, time line

: Printing of scale, channel No. and tag (only if

external (disable) is selectable. Reference junction compensation accuracy:

they are programmed) with every specific interval gChart speed : Printing of chart speed with every specific

K, E, J, T, N and Platinel II…±0.5°C or less Other than the above*… ±0.1°C or less *Except PtRh40-PtRh20 Input resolution: Approx. 1/56000 (converted into reference

: Printing of time, channel No., alarm type and

alarm level when an alarm activates. Printing time, channel No. and alarm level when the alarm is reset. Memory volume is max. 48 data iProgramming

range) Burnout: Effective input… Thermocouple, resistance thermometer Programming range …Selection from up-scale, down-scale or disable for each channel Allowable signal source resistance:

: Marking when changed parameters are

stored j POC markMarking when time-axis synchronization is

DC voltage inputs, thermocouple inputs

programmed k List printingPrinting of parameters (list 1 or 2) on demand

…1kΩ or less* Resistance thermometer inputs

List 1Chart speed, range/printing range,

…10Ω or less * Without burnout Input resistance: Thermocouple input …About 8MΩ

scale, subtract printing, periodic data printing, printing format and alarm List 2: Chart speed (3 speeds), alarm output,

DC voltage inputs (range of ±2 V or less)

time-axis synchronization, remote contacts and math expression List 3: All parameters (List 1 + List 2)

…About 8MΩ DC voltage inputs (range of ±5 V or less)

Show 14 more labels / short text on this page
Maximum input voltage · DC voltage · inputs, thermocouple · NiMo-Ni, CR-AuFe, PtRh40-PtRh20 · WRe5-WRe26, W-WRe26, · condition) · Thermo- · couple · scale, tag ゙ · interval hAlarm · Activation/ Reset · change mark · …About 1MΩ · -126-
Page 131 of the Omega RD200 user manual — 3 Indication and Display
Page 131 · click to enlarge
Page 131

22. SPECIFICATIONS

4 Alarm

l MessagePrinting of message in message No. selected
Numbers of messageMax. 5 Message programming: Max. 15 digits Chart paper: Depending on models

Alarm point: Numbers of channel x numbers of level (4) Alarm display: Alarm status lamp lights and the

measured value blinks at alarm activated channel. Alarm display screen indicates alarm channel and alarm type at level digit. Alarm type: Selectable from the following 10 types for each

h RD200 Folding type 100mm 114mm 10m RD2800 Folding type 180mm 200mm 20m Chart speed: 1 to 600mm/h, 1 to 200mm/h

H: High limit L: Low limit Absolute

Default … RD20020mm/h

value E: High limit with standby F: Low limit with standby Rate-of- change *1 U: Increase limit D: Decrease limit

RD280025mm/h Chart speed accuracy: Within ± 0.1%, For the chart scale

after feeding 1000mm or more. Skip function: No display or printing is executed for the

B: Differential high limit S: Differential low limit Difference

channels for which no range is programmed. Time axis synchronization:

*2 JDifferential high limit

K: Differential low limit

with standby *1: Variation range per unit time (Note) (Note) Measuring cycle x measuring count (1 to 20) *2: Difference between two channels Alarm deadband: 0.1 to 9.9% of scale programming range Alarm output: Option

On or off selectable with key operation Subtract printing:

Printing of difference between measured value and reference channel (or reference value) Pen lift: Pens are automatically lifted up when printing is off.

3 Indication and Display

5 Programming and Operation

Trace printing indicationBargraph display for each channel

Model No. or segment Channel color segment

Key types: See Section 7.1. Key functions: See Section 7.2 . Basic operation: Basic operation with keys

RD200 51 5-segment each RD2800 101 10-segment each Digital display: 16- and 7-segment LCD with cursor

c Printing ON/OFF d Chart paper feeding e Operation screen switching Operation: Operation with keys

16-segment … Character height: 7.5 mm, orange 7-segment …. Character height: 6.5 mm, white Model 16-segment 7-segment RD200 1 digit 20 digits RD2800 1 digit 30 digits Display items: cSimultaneous display of measured values

Selection of c Digital data printing *1 d List printing *1 e Message printing *1 and f Printing format*2 *1: Operation with remote contacts (option) is also available *2: Only available with option Pen change mode: Pen moves to the position to beeasily

for all channels d Year/month/day (only RD200) e Time f Chart speed g Alarm status Display switching: Each time is pressed, the

changed. Standard programming:

display item changes.

RD200 .… c → def → g RD2800 … cef → g Measured value display: -9999 to 99999 for each channel

Range/printing range, °C/°F computation, chart speed, time, scale, skip, subtract printing, alarm, alarm deadband, periodic time printing, engineering unit, tag, message, burnout, passcode/key lock, input filter Optional programming:

(Decimal point can be placed at desired position with the scale programming ) Status: Following 6 status lamps are available.

Alarm output…Relay No., AND/OR, output mode Remote contacts…Terminal allocation for operation, 3

Omega RD200 — 3 Indication and Display diagram (page 131)
Fig. 251 · 3 Indication and Display

Display text Condition to light Color RECORD ON Printing ON Green KEY LOCK Key lock Blue ALARM Alarm occurrence Red CHART END Just before chart paper ends Red

chart speeds, operation record position Printing format……Automatic range-shift printing,

compressed/expanded, zone printing Communications interface…Communications protocol,

communications specifications Math expression…Selection of types, parameters Totalization…Starting time, interval Engineering port: All parameters can be programmed with

POC Time axis sync. ON Blue Chart illumination: CFL

engineering software “PASS” (option) installed on a personal computer

Show 7 more labels / short text on this page
Total lengt · Model System Printing · Total width · width · alarm point · with standby · -127-
Page 132 of the Omega RD200 user manual — 7 General specifications
Page 132 · click to enlarge
Page 132

22. SPECIFICATIONS

ColorDoor … Frame: Black (equivalent to Munsell N3.0)

6 Operation Conditions

Front plateTransparent Enclosure…Gray (equivalent to Munsell N7.0) Mounting: Panel mounting Terminal screws: Power terminals …M4.0
Operation conditionsRefer to the following table

Terminals other than above …M3.5 Weight: RD200 …Approx. 4.0 kg (with fully options) RD2800 …Approx. 9.0 kg (with fully options) Dimension: RD200…144×144×258 (mm)

Ambient temperature 21 to 25°C 0 to 31°C 31 to 50°C*

Max. ambient

humidity 80%RH 80%RH 80 to 50%RH*

RD2800…288×288×220 (mm) Panel cutout:RD200…138×138 (mm)

Min. ambient

RD2800…281×281 (mm) Clock accuracy: Within ±2 minutes per 30 day

voltage 100VAC ± 1% 90 to 264VAC

(Under reference operating conditions, Except errors when power supply is turned on or off) Packing material: Layered cardboard is used as cushion

frequency 50/60Hz ± 0.5% 50/60Hz ± 2%

Left/right 0° 0 to 10° Forward

tilting 0° 0° Posture

tilting 0° 0 to 30°

8 Safety Regulations

Warming-up time 30 min. or more Not necessary *Decrease from 80% RH at 31°C, and decrease to 50% at 40°C Transportation condition:

CE: Conforms to EMC and low voltage directives UL: UL3111-1 (Approval pending) CSA (C-UL): CSA …C22.2 No.1010 (Approval pending) IP: Conforms to IEC529 IP54

Ambient temperature/humidity… -20 to 60°C, 5 to 90%RH

(without condensation) Vibration…10 to 60Hz, 4.9 m/S2* Impact…392 m/S2* * When shipping out from a factory Storage condition:

Ambient temperature/humidity… -20 to 60°C, 5 to 90%RH

7 General specifications

Rated power voltage: 100 to 240 VAC, 50/60 Hz Power consumption: Maximum 60V A Power failure protection: Programmed parameters stored

into EEPROM memory. Clock circuit sustained for minimum 8 years by a lithium battery. (8 hours operation per day) Insulation resistance:

Between primary and protective conductor terminals

…20MΩ or more at 500 V DC Between secondary and protective conductor terminals

…20MΩ or more at 500 V DC Between primary and secondary terminals

…20MΩ or more at 500 V DC Dielectric strength:

Between primary and protective conductor terminals

…1 minute at 1500 V AC Between secondary and protective conductor terminals

…1 minute at 500 V AC Between primary and secondary terminals

…1 minute at 2300 V AC Primary terminals: Power terminals, Alarm output terminals

(MOS relay, mechanical relay “a” contact) Secondary terminals: Measuring input terminals, Remote

contacts terminals, Communications interface terminals , Alarm output terminals (Mechanical relay “c” contact), Case assembly material: Door Frame …ABS resin

Show 10 more labels / short text on this page
Items Reference · operation Normal operation · humidity 20%RH 20%RH · Power supply · Power supply · material · Backward · (without condensation) · Front plate…Polycarbonate Enclosure…Steel · -128-
Page 133 of the Omega RD200 user manual — DC
Page 133 · click to enlarge
Page 133

22. SPECIFICATIONS

9 Accuracy Ratings

No. Input type Measuring range Reference

No. Input type Measuring range Reference

rating 01 -13.80 to 13.80mV ±13.8mV 02 -27.60 to 27.60mV ±27.6mV 03 -69.00 to 69.00mV ±69.0mV 04 -200.0 to 200.0mV ±200.0mV 05

rating 79 Pt50 -200.0 to 649.0°C 220Ω ±0.1% ± 1 d 80 Pt-Co 4.0 to 374.0K 220Ω ±0.15% ± 1 d Note) “± 1 d” at accuracy rating stands for ± 1 digit. Note) Thermocouple inputs do not include reference junction

-500.0 to 500.0mV ±500.0mV 06 -2.000 to 2.000V ±2V 07 -5.000 to 5.000V ±5V 08 -10.00 to 10.00V ±10V 09 -20.00 to 20.00V ±20V 10

compensation accuracy. The Indication equivalent to 200 µV may vary under the test environment of EMC directive. Exceptions to accuracy rating

±0.1% ± 1 d

DC

(V)

-50.00 to 50.00V ±50V

Input type Measuring range Accuracy rating K, E, J, T, L -200 to 0°C R, S 0 to 400°C ±0.2% ± 1 digit

21 -200.0 to 300.0°C ±13.8mV 22 -200.0 to 600.0°C ±27.6mV 23 K

-200.0 to 1370 °C ±69.0mV 24 -200.0 to 200.0°C ±13.8mV 25 -200.0 to 350.0°C ±27.6mV 26 E

0 to 400°C Not specified B 400 to 800°C ±0.15% ± 1 digit N, U -200 to 0°C ±0.3% ± 1 digit 0 to 100°C ±4% ± 1 digit W-WRe26 100 to 300°C ±0.5 % ± 1 digit 0 to 300°C ±1.5% ± 1 digit PtRh40-PtRh20 300 to 800°C ±0.8% ± 1 digit 0 to 20K ±0.5% ± 1 digit CR-AuFe 20 to 50K ±0.3% ± 1 digit Pt100 (’97) 700 to 850°C ±0.15% ± 1 digit Pt-Co 4 to 50K ±0.3% ± 1 digit Note) Thermocouple input is converted into the reference

-200 to 900°C ±69.0mV 27 -200.0 to 250.0°C ±13.8mV 28 -200.0 to 500.0°C ±27.6mV 29 J

±0.1% ± 1 d

-200 to 1200°C ±69.0mV 30 -200.0 to 250.0°C ±13.8mV 31 T -200.0 to 400.0°C ±27.6mV 32 0 to 1200 °C ±13.8mV 33 R 0 to 1760 °C ±27.6mV 34 0 to 1300 °C ±13.8mV 35 S 0 to 1760 °C ±27.6mV 36 B 0 to 1820 °C ±13.8mV

range.

37 -200.0 to 400.0°C ±13.8mV 38 -200.0 to 750.0°C ±27.6mV 39 N

±0.15% ± 1 d

-200 to 1300 °C ±69.0mV 40 W・WRe26 0 to 2315°C ±69.0mV

41 WRe5・ WRe26 0 to 2315°C ±69.0mV

43 PtRh40- PtRh20 0 to 1888°C ±13.8mV

44 -50.0 to 290.0°C ±13.8mV 45 -50.0 to 600.0°C ±27.6mV 46 NiMo-Ni

±0.2% ± 1 d

-50 to 1310 °C ±69.0mV 47 CR-AuFe 0.0 to 280.0K ±13.8mV

48 0.0 to 350.0°C ±13.8mV 49 0.0 to 650.0°C ±27.6mV 50 Platinel ll

0 to 1395°C ±69.0mV 51 -200.0 to 250.0°C ±13.8mV 52 -200.0 to 500.0°C ±27.6mV 53 U

±0.15% ± 1 d

-200.0 to 600.0°C ±69.0mV

54 -200.0 to 250.0°C ±13.8mV 55 -200.0 to 500.0°C ±27.6mV 56 L

±0.1% ± 1 d

-200 to 900°C ±69.0mV

70 -140.0 to150.0°C 160Ω ±0.15% ± 1 d 71 -200.0 to 300.0°C 220Ω 72

Pt100 (JIS’97) -200.0 to 850.0°C 400Ω ±0.1% ± 1 d

73 -140.0 to 150.0°C 160Ω ±0.15% ± 1 d 74 -200.0 to 300.0°C 220Ω 75

JIS’89 -200.0 to 649.0°C 400Ω ±0.1% ± 1 d

76 -140.0 to 150.0°C 160Ω ±0.15% ± 1 d 77 -200.0 to 300.0°C 220Ω 78 JPt100

-200.0 to 649.0°C 400Ω ±0.1% ± 1 d

Show 7 more labels / short text on this page
Accuracy · Accuracy · range · range · DC (mV) · Pt100 QPt100 · -129-
Page 134 of the Omega RD200 user manual — 10 Option
Page 134 · click to enlarge
Page 134

22. SPECIFICATIONS

3. Printing format Trace printing format, c Automatic range shift, d Compressed/Expanded or e Zone scale, is selectable for each channel.. The format e is common to all channels.

10 Option

1. Alarm outputs Output types: c Alarm point output d FAIL output e Chart

paper end (C.End) output Numbers of outputs: RD200 … 6 points

RD2800 …6 or 12 points Contact capacity (with resistive load):

Printing area can be divided up to 5 ranges. It switches the range according to the measured values automatically for printing. Compressed/

cMOS relay output … 240 V (AC, DC), 50 mA (AC, DC) d Mechanical relay (Common for “a” and “c” contacts) …

Specific printing range is compressed or expanded for printing.

100VAC 0.5 A, 240VAC 0. 2A, 100VDC 0.3 A Output wiring: AND/OR output selection is available for each

Printing area is divided into 2 to 4 zones*. Data of each channel is printed in the relevant zone. *RD200 … 2, RD2800 … 2 to 4 4) Others Options Detail Communications

alarm points (including FAIL and C.End) Relay coil phase: Energize/Not energize selectable* Relay output latch: Hold/Not hold selectable* Alarm display latch: Hold/Not hold selectable*

*Common to all relays Dimension: The following length will be added for the depth.

RD200…14 mm (MOS relay, Mechanical “c”

Specify one of the following modes. RS-232C, RS-422A, RS-485

contact) 25 mm (Mechanical “a” contact) RD2800…16 mm (MOS relay, Mechanical “c”

Capable up to ±50mA DC by connecting an input resistance to the measuring input terminal. 100Ω: Applicable current ±50mA 250Ω: Applicable current ±20mA

Input resistance for current input*

27 mm (Mechanical “a” contact) 2) Remote contacts Input terminals: 4 points (Common terminals: 2 points) Input signal: Contact (Open/Shorted) Voltage when the contact is open: Approx. 5 V Current when the contact is shorted: Approx. 2 mA Operations: Up to 4 input terminals can be selectable in the

One of following math expressions is selectable for each channel. Addition, subtraction, multiplication, division, natural logarithm, logarithm, exponential, square root, temperature/humidity, maximum, minimum, average value, absolute value, data communications input*, flow rate compensation. *This function is only available for the “communications interface” (option).

following table.

Operation Terminal c Printing ON/OFF, 3 chart speeds 2 d Message (No.1 to No.5) printing 4 e Message (No. 1 and No. 2) printing 2 f Digital data printing 1 g List 1 printing 1 h List 2 printing 1 i List 3 printing 1 j Operation record A 1 k Operation record B 1 l Operation record C 1 Operation record D 1 Reset of totalization* 1 *This operation is only available for the “totalization” (option).

Totalizing measured values and calculation results. Reset function is executed with programmed intervals or remote contacts signals*. *This function is only available for the “remote contacts” (option). Transmitter power supply*

Power supply unit for the transmitter generating input signal to the recorder

16 m chart paper RD200 only …. Total chart paper length of 15.6 m * Exclusive instruction manual is provided.

Show 8 more labels / short text on this page
Automatic range-shift · Expanded · Zone scale · interface* · contact) · Math expression* · Totalizing* · -130-
Fix-it

Troubleshooting

Common problems, likely causes, and what to do.

ProblemPossible causeRemedy
1) Measured values are unstable.• Check if the measuring terminals are loose. • Check if the input signal is unstable. • Check if the thermocouple is connected with another instrument in parallel and the burnout function is programmed to [Enable]. p.127
2) Check if the measured value is displayed as shown below. • Check if the input terminals are connected correctly. • Check if the input terminals are loose. • Check if the input signal wires are disconnected. • Check if the input signal exceeds the measuring range. p.127
3) An error occurs.• Check if the selection of ºC/ºF computation is correct. • Check if the input signal is correct. • Check if an extension wire is connected to the input terminal. (Thermocouple input type only) • Check the scale and adjust the measured values if any error is detected. (See Section 18.1.) p.127
4) Influences by ambient temperature (Thermocouple input type only)• Check if the RJ selection is 0 (disable) in [Range/Printing range] programming. (This is negligible if reference junction compensation is executed externally.) • Check if the terminal cover is mounted. p.127
1) Measured value blinks.Measured value of the channel in which an alarm activates. See Section 10.3. p.128
2) Measured value is blank.No measured value is displayed for the skipped channel. p.128
3) Year, month, day, hour and minute are displayed.This is one of the operation screens (Clock display). See Section 6.4. p.128
4) “” is displayed at the left end.This is one of the operation screens (Alarm activation screen). See Section 6.4. p.128
1) No printing is executed.• Check if the printer is turned off. See Section 6.2. • Check if the cassette ribbon is mounted normally. p.128
2) Digital printing is not executed.All digital printings are not executed if the chart speed is programmed to be faster than 150mm/h. See Section 10.2. p.128
3) Printing color is pale.Replace the cartridge pen or the plotter pen. p.128
4) Chart paper feed is abnormal.• Check if the chart paper is unlatched from the sprocket. • Check if the chart paper cassette is securely inserted into the internal unit. • Check if the chart paper has been shuffled before loading it. p.128
1) Keys are not acceptable.Keys are locked. Unlock them. See Section 11.12. p.128
2) Parameters cannot be changed. p.128
3) Prarameters remain unchanged after changing them.To store parameters to be programmed for each channel, press SHIFT and S H I F T simultaneously. If not, parameters are not stored into memory. See Section 8.2 and other sections for programming. p.128

Ask about the Omega RD200

Owners and our team usually reply within a day — and your question helps the next person too.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
No questions yet about the Omega RD200

Be the first to ask — other owners or our team will answer.

Original document

Omega RD200 manual PDF

134 pages·2.6 MB·Sensor·PDF
Click any page above to open it here
Page 1 / …
100%
Loading pages…