Instruction/ maintenance manual of the product FTB600B Omega
Go to page of 26
omega.com e-mail: info@omega.com For latest product manuals: omegamanual.info FTB600B Series Ultra-Low Flow Sensors Shop online at U ser ’ s Guide.
OMEGAnet ® On-Line Service Internet e-mail omega.com info@omega.com It is the policy of OMEGA Engineering, Inc. to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives.
TABLE OF CONTENTS FTB600B Series Ultra-Low Flow Sensors i Section Page Section 1 Introduction ...................................................................... 1 A. General Description .................................................... 1 B. Principle of Operation .
ii FTB600B Series Ultra-Low Flow Sensors NOTES:.
1 1 FTB600B Series Installation IMPORTANT: READ INSTRUCTIONS THOROUGHLY BEFORE INSTALLING FLOW METER Section 1 - Introduction A. General Description The FTB600B Series is an axial paddle wheel turbine type flow meter based on the pelton wheel principle.
2 FTB600B Series Installation 1 C. Material Characteristics of PVDF (Polyvinylidene Fluoride) 1. Material of construction Trade name - Kynar All wetted parts of the FTB600B Series are PVDF, excluding the O-ring. Wetted parts include any part of the meter that will or could come in contact with the fluid.
3 2 FTB600B Series Operating Parameters Section 2. - Operating Parameters A. Temperature Since the FTB600B Series has printed circuit boards molded into the body of the meter it is strongly recommended that 180°F not be exceeded. Exceeding 180°F can cause irreparable damage to the circuit boards.
4 FTB600B Series Operation Parameters 2.
5 FTB600B Series Operating Parameters 2 C. Recommended Viscosity Range 1-5 cSt (w/o correction) The effects of changing viscosity on the FTB600B Series are the same as any other turbine flow meter.
6 3 FTB600B Series Silicone Treatment Section 3 - Silicone Treatment Silicone treatment is standard for all types of the FTB600B Series series electronics.
7 4 FTB600B Series Infrared Sensors Section 4 - Infrared Sensor A. Supply Voltage 24 Vdc. Do not exceed 24 Vdc. Doing so can cause overheating and eventual failure of all PC boards. Printed circuit boards are non- repairable. B. Frequency Output 1. Square wave pulse, unscaled (See Figure 4-1) 2.
8 FTB600B Series Infrared Sensors 4 C. Frequency Ranges for (non-cartridge models) pulses/l Model Frequency (Hz) K-Factor FTB601B 60 to 1200 36,000 FTB602B 40 to 1200 8,000 FTB603B 27 to 800 3,200 FTB604B 20 to 600 1,200 FTB605B 18.75 to 562 450 FTB606B 15 to 450 225 D.
9 FTB600B Series Infrared Sensors 4 Figure 4-1. Output Frequency Waveforms.
10 5 FTB600B Series Installation Section 5 - Installing the FTB600B Series in the fluid line. A. Make sure the fluid is compatible with PVDF (polyvinylidene fluoride) and meets viscosity, pressure and temperature parameters of the FTB600B Series. The fluid must also meet filtration requirements as listed in 3D.
11 FTB600B Series Installation 5 Figure 5-1. Supply Voltage and Signal Output Connections.
12 FTB600B Series Installation 5 Table 5-1. Piping Table.
13 FTB600B Series Installation 5 Table 5-1. Piping Table Con't..
14 6 FTB600B Series Specifications Section 6 - Specifications Accuracy: ±3% of reading Repeatability: ±0.1% of reading Linearity: ±1% of reading Viscosity Range: 1 to 15 centistokes Working Pressur.
15 7 FTB600B Series Pressure Drop Curves Section 7 - Pressure Drop Curves Pressure Drop FTB600B Series Liquid Turbine - Flowmeter FTB601 50 0.5 0 1.0 Liters/minute kP a 1.
16 FTB600B Series Pressure Drop Curves 7 Pressure Drop FTB600B Series Liquid Turbine - Flowmeter FTB602 250 2.0 0 4.0 Liters/minute kP a 8 .0 6.0 10.0 200 150 100 50 0.
17 FTB600B Series Pressure Drop Curves 7 Pressure Drop FTB600B Series Liquid Turbine - Flowmeter FTB603 200 51 0 1 5 0 Liters/minute kP a 160 1 8 0 140 100 8 0 60 40 20 120 0.
18 FTB600B Series Pressure Drop Curves 7 Pressure Drop FTB600B Series Liquid Turbine - Flowmeter FTB604 200 10 20 30 0 Liters/minute kP a 160 1 8 0 140 100 8 0 60 40 20 120 0.
19 FTB600B Series Pressure Drop Curves 7 Pressure Drop FTB600B Series Liquid Turbine - Flowmeter FTB605A 160 140 120 100 8 0 60 40 20 10 20 30 40 50 60 0 Liters/minute kP a 0.
20 FTB600B Series Pressure Drop Curves 7 Pressure Drop FTB600B Series Liquid Turbine - Flowmeter FTB606A 1 8 0 160 140 120 100 8 0 60 40 20 10 20 30 40 50 60 70 8 09 0 0 Liters/minute kP a 0.
FOR WARRANTY RETURNS, please have the following information available BEFORE contacting OMEGA: 1. Purchase Order number under which the product was PURCHASED, 2. Model and serial number of the product under warranty, and 3. Repair instructions and/or specific problems relative to the product.
TEMPERATURE 䡺 ⻬ Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies 䡺 ⻬ Wire: Thermocouple, RTD & Thermistor 䡺 ⻬ Calibrators & Ice Point References 䡺 .
An important point after buying a device Omega FTB600B (or even before the purchase) is to read its user manual. We should do this for several simple reasons:
If you have not bought Omega FTB600B yet, this is a good time to familiarize yourself with the basic data on the product. First of all view first pages of the manual, you can find above. You should find there the most important technical data Omega FTB600B - thus you can check whether the hardware meets your expectations. When delving into next pages of the user manual, Omega FTB600B you will learn all the available features of the product, as well as information on its operation. The information that you get Omega FTB600B will certainly help you make a decision on the purchase.
If you already are a holder of Omega FTB600B, but have not read the manual yet, you should do it for the reasons described above. You will learn then if you properly used the available features, and whether you have not made any mistakes, which can shorten the lifetime Omega FTB600B.
However, one of the most important roles played by the user manual is to help in solving problems with Omega FTB600B. Almost always you will find there Troubleshooting, which are the most frequently occurring failures and malfunctions of the device Omega FTB600B along with tips on how to solve them. Even if you fail to solve the problem, the manual will show you a further procedure – contact to the customer service center or the nearest service center